Introduction: von Willebrand disease (VWD) diagnosis starts with first level tests: factor VIII coagulant activity, VWF antigen (VWF:Ag) and platelet-dependent VWF activity (VWF:RCo, VWF:Ab, VWF:GPIbR or VWF:GPIbM). The VWF collagen binding (VWF:CB) assay measures the binding capacity of von Willebrand factor (VWF) to collagen. Aim: To assess, in previously diagnosed VWD patients, the performance of a fully automated chemiluminescent test panel including VWF:Ag, VWF:GPIbR and VWF:CB assays. Methods: The patients, historically evaluated using in-house VWF:Ag and VWF:CB assays and an automated latex enhanced immunoassay VWF:GPIbR method, were re-evaluated using the VWF test panel HemosIL AcuStar. Results: The VWF:GPIbR/VWF:Ag and VWF:CB/VWF:Ag obtained by means of AcuStar showed an overall good concordance with the corresponding data obtained at the time of the historical diagnosis. When discrepancies occurred, these were generally due to the lower VWF:CB/VWF:Ag obtained with AcuStar as compared with that obtained with the historical methods and this affected particularly the diagnosis of VWD type 2M. Together, the AcuStar VWF:GPIbR/VWF:Ag and VWF:CB/VWF:Ag were able to distinguish type 1 from types 2A, 2B and 2M, whereas no distinction was possible between type 2A and 2B. Conclusion: The AcuStar panel offers a good performance in the differential diagnosis between VWD type 1 and 2A/2B patients. A high rate of coincidence with historical diagnosis was obtained for VWD types 3, 2A/2B and 1. Even though in some cases more tests (eg, RIPA/multimeric analysis) are needed to complete an accurate VWD classification, the AcuStar panel is considered a sensitive, rapid and reliable tool to diagnose VWD patients.

Evaluation of a fully automated von Willebrand factor assay panel for the diagnosis of von Willebrand disease / F. Stufano, L. Baronciani, P. Bucciarelli, M. Boscarino, P. Colpani, M.T. Pagliari, F. Peyvandi. - In: HAEMOPHILIA. - ISSN 1351-8216. - 26:2(2020 Mar), pp. 298-305.

Evaluation of a fully automated von Willebrand factor assay panel for the diagnosis of von Willebrand disease

L. Baronciani;P. Bucciarelli;M.T. Pagliari;F. Peyvandi
2020

Abstract

Introduction: von Willebrand disease (VWD) diagnosis starts with first level tests: factor VIII coagulant activity, VWF antigen (VWF:Ag) and platelet-dependent VWF activity (VWF:RCo, VWF:Ab, VWF:GPIbR or VWF:GPIbM). The VWF collagen binding (VWF:CB) assay measures the binding capacity of von Willebrand factor (VWF) to collagen. Aim: To assess, in previously diagnosed VWD patients, the performance of a fully automated chemiluminescent test panel including VWF:Ag, VWF:GPIbR and VWF:CB assays. Methods: The patients, historically evaluated using in-house VWF:Ag and VWF:CB assays and an automated latex enhanced immunoassay VWF:GPIbR method, were re-evaluated using the VWF test panel HemosIL AcuStar. Results: The VWF:GPIbR/VWF:Ag and VWF:CB/VWF:Ag obtained by means of AcuStar showed an overall good concordance with the corresponding data obtained at the time of the historical diagnosis. When discrepancies occurred, these were generally due to the lower VWF:CB/VWF:Ag obtained with AcuStar as compared with that obtained with the historical methods and this affected particularly the diagnosis of VWD type 2M. Together, the AcuStar VWF:GPIbR/VWF:Ag and VWF:CB/VWF:Ag were able to distinguish type 1 from types 2A, 2B and 2M, whereas no distinction was possible between type 2A and 2B. Conclusion: The AcuStar panel offers a good performance in the differential diagnosis between VWD type 1 and 2A/2B patients. A high rate of coincidence with historical diagnosis was obtained for VWD types 3, 2A/2B and 1. Even though in some cases more tests (eg, RIPA/multimeric analysis) are needed to complete an accurate VWD classification, the AcuStar panel is considered a sensitive, rapid and reliable tool to diagnose VWD patients.
burden of illness; chronic illness; collagen binding assay; laboratory diagnosis; von Willebrand disease; von Willebrand factor deficiency
Settore MED/09 - Medicina Interna
mar-2020
Article (author)
File in questo prodotto:
File Dimensione Formato  
hae.13929.pdf

accesso riservato

Tipologia: Publisher's version/PDF
Dimensione 741.33 kB
Formato Adobe PDF
741.33 kB Adobe PDF   Visualizza/Apri   Richiedi una copia
Pubblicazioni consigliate

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/2434/719604
Citazioni
  • ???jsp.display-item.citation.pmc??? 2
  • Scopus 7
  • ???jsp.display-item.citation.isi??? 7
social impact