Thrombogenic Risk Assessment of Transcatheter Prosthetic Heart Valves Using a Fluid-Structure Interaction Approach
Authors:
Kyle Baylous,
Brandon Kovarovic,
Salwa Anam,
Ryan Helbock,
Marvin Slepian,
Danny Bluestein
Abstract:
Prosthetic heart valve interventions such as TAVR have surged over the past decade, but the associated complication of long-term, life-threatening thrombotic events continues to undermine patient outcomes. Thus, improving thrombogenic risk analysis of TAVR devices is crucial. In vitro studies for thrombogenicity are typically difficult to perform. However, revised ISO testing standards include com…
▽ More
Prosthetic heart valve interventions such as TAVR have surged over the past decade, but the associated complication of long-term, life-threatening thrombotic events continues to undermine patient outcomes. Thus, improving thrombogenic risk analysis of TAVR devices is crucial. In vitro studies for thrombogenicity are typically difficult to perform. However, revised ISO testing standards include computational testing for thrombogenic risk assessment of cardiovascular implants. We present a fluid-structure interaction (FSI) approach for assessing thrombogenic risk of prosthetic heart valves.
An FSI framework was implemented via the incompressible computational fluid dynamics multi-physics solver of the Ansys LS-DYNA software. The numerical modeling approach for flow analysis was validated by comparing the derived flow rate of the 29-mm CoreValve device from benchtop testing and orifice areas of commercial TAVR valves in the literature to in silico results. Thrombogenic risk was analyzed by computing stress accumulation (SA) on virtual platelets seeded in the flow fields via Ansys EnSight. The integrated FSI-thrombogenicity methodology was subsequently employed to examine hemodynamics and thrombogenic risk of TAVR devices with two approaches: 1) engineering optimization and 2) clinical assessment.
Our methodology can be used to improve the thromboresistance of prosthetic valves from the initial design stage to the clinic. It allows for unparalleled optimization of devices, uncovering key TAVR leaflet design parameters that can be used to mitigate thrombogenic risk, in addition to patient-specific modeling to evaluate device performance. This work demonstrates the utility of advanced in silico analysis of TAVR devices that can be utilized for thrombogenic risk assessment of other blood recirculating devices.
△ Less
Submitted 17 June, 2024;
originally announced June 2024.
In Silico Fatigue Optimization of TAVR Stent Designs with Physiological Motion in a Beating Heart Model
Authors:
Kyle Baylous,
Ryan Helbock,
Brandon Kovarovic,
Salwa Anam,
Marvin Slepian,
Danny Bluestein
Abstract:
The rapid expansion of TAVR to younger, low-risk patients raises concerns regarding device durability. Necessarily, extended stent lifetime will become more critical for new generation devices. In vitro methods commonly used for TAVR stent fatigue testing exclude the effects of the beating heart. We present a more realistic in silico stent fatigue analysis utilizing a beating heart model in which…
▽ More
The rapid expansion of TAVR to younger, low-risk patients raises concerns regarding device durability. Necessarily, extended stent lifetime will become more critical for new generation devices. In vitro methods commonly used for TAVR stent fatigue testing exclude the effects of the beating heart. We present a more realistic in silico stent fatigue analysis utilizing a beating heart model in which TAVR stents experience complex, nonuniform dynamic loading. Virtual TAVR deployments were simulated in the SIMULIA Living Heart Human Model of a beating heart using stent models of the self-expandable nitinol 26-mm CoreValve and Evolut R devices, and a 27-mm PolyV-2. Stent deformation was monitored over three cardiac cycles, and fatigue resistance was evaluated for the nitinol stents using finite element analysis via ABAQUS/Explicit. In all models, there were elements in which strains exceeded fatigue failure. The PolyV-2 stent had far fewer failing elements since its struts were optimized to reduce the strain in stent joints, achieving better fatigue resistance in the stent crown and waist elements. Different stent sections showed markedly different fatigue resistance due to the varying loading conditions. This study demonstrates the utility of advanced in silico analysis of devices deployed within a beating heart that mimics in vivo loading, offering a cost-effective alternative to human or animal trials and establishing a platform to assess the impact of device design on device durability. The limited fatigue life of TAVR stents indicated here highlights a clinical complication that may eventually develop as younger, lower-risk TAVR patients, age.
△ Less
Submitted 13 September, 2023;
originally announced September 2023.