The Design, Construction, and Commissioning of the KATRIN Experiment
Authors:
M. Aker,
K. Altenmüller,
J. F. Amsbaugh,
M. Arenz,
M. Babutzka,
J. Bast,
S. Bauer,
H. Bechtler,
M. Beck,
A. Beglarian,
J. Behrens,
B. Bender,
R. Berendes,
A. Berlev,
U. Besserer,
C. Bettin,
B. Bieringer,
K. Blaum,
F. Block,
S. Bobien,
J. Bohn,
K. Bokeloh,
H. Bolz,
B. Bornschein,
L. Bornschein
, et al. (204 additional authors not shown)
Abstract:
The KArlsruhe TRItium Neutrino (KATRIN) experiment, which aims to make a direct and model-independent determination of the absolute neutrino mass scale, is a complex experiment with many components. More than 15 years ago, we published a technical design report (TDR) [https://publikationen.bibliothek.kit.edu/270060419] to describe the hardware design and requirements to achieve our sensitivity goa…
▽ More
The KArlsruhe TRItium Neutrino (KATRIN) experiment, which aims to make a direct and model-independent determination of the absolute neutrino mass scale, is a complex experiment with many components. More than 15 years ago, we published a technical design report (TDR) [https://publikationen.bibliothek.kit.edu/270060419] to describe the hardware design and requirements to achieve our sensitivity goal of 0.2 eV at 90% C.L. on the neutrino mass. Since then there has been considerable progress, culminating in the publication of first neutrino mass results with the entire beamline operating [arXiv:1909.06048]. In this paper, we document the current state of all completed beamline components (as of the first neutrino mass measurement campaign), demonstrate our ability to reliably and stably control them over long times, and present details on their respective commissioning campaigns.
△ Less
Submitted 11 June, 2021; v1 submitted 5 March, 2021;
originally announced March 2021.
-
An improved upper limit on the neutrino mass from a direct kinematic method by KATRIN
Authors:
M. Aker,
K. Altenmüller,
M. Arenz,
M. Babutzka,
J. Barrett,
S. Bauer,
M. Beck,
A. Beglarian,
J. Behrens,
T. Bergmann,
U. Besserer,
K. Blaum,
F. Block,
S. Bobien,
K. Bokeloh,
J. Bonn,
B. Bornschein,
L. Bornschein,
H. Bouquet,
T. Brunst,
T. S. Caldwell,
L. La Cascio,
S. Chilingaryan,
W. Choi,
T. J. Corona
, et al. (184 additional authors not shown)
Abstract:
We report on the neutrino mass measurement result from the first four-week science run of the Karlsruhe Tritium Neutrino experiment KATRIN in spring 2019. Beta-decay electrons from a high-purity gaseous molecular tritium source are energy analyzed by a high-resolution MAC-E filter. A fit of the integrated electron spectrum over a narrow interval around the kinematic endpoint at 18.57 keV gives an…
▽ More
We report on the neutrino mass measurement result from the first four-week science run of the Karlsruhe Tritium Neutrino experiment KATRIN in spring 2019. Beta-decay electrons from a high-purity gaseous molecular tritium source are energy analyzed by a high-resolution MAC-E filter. A fit of the integrated electron spectrum over a narrow interval around the kinematic endpoint at 18.57 keV gives an effective neutrino mass square value of $(-1.0^{+0.9}_{-1.1})$ eV$^2$. From this we derive an upper limit of 1.1 eV (90$\%$ confidence level) on the absolute mass scale of neutrinos. This value coincides with the KATRIN sensitivity. It improves upon previous mass limits from kinematic measurements by almost a factor of two and provides model-independent input to cosmological studies of structure formation.
△ Less
Submitted 13 September, 2019;
originally announced September 2019.
-
Neutron-induced background in the CONUS experiment
Authors:
J. Hakenmüller,
C. Buck,
K. Fülber,
G. Heusser,
T. Klages,
M. Lindner,
A. Lücke,
W. Maneschg,
M. Reginatto,
T. Rink,
T. Schierhuber,
D. Solasse,
H. Strecker,
R. Wink,
M. Zboril,
A. Zimbal
Abstract:
CONUS is a novel experiment aiming at detecting elastic neutrino nucleus scattering in the fully coherent regime using high-purity Germanium (Ge) detectors and a reactor as antineutrino ($\barν$) source. The detector setup is installed at the commercial nuclear power plant in Brokdorf, Germany, at a very small distance to the reactor core in order to guarantee a high flux of more than 10…
▽ More
CONUS is a novel experiment aiming at detecting elastic neutrino nucleus scattering in the fully coherent regime using high-purity Germanium (Ge) detectors and a reactor as antineutrino ($\barν$) source. The detector setup is installed at the commercial nuclear power plant in Brokdorf, Germany, at a very small distance to the reactor core in order to guarantee a high flux of more than 10$^{13}\barν$/(s$\cdot$cm$^2$). For the experiment, a good understanding of neutron-induced background events is required, as the neutron recoil signals can mimic the predicted neutrino interactions. Especially neutron-induced events correlated with the thermal power generation are troublesome for CONUS. On-site measurements revealed the presence of a thermal power correlated, highly thermalized neutron field with a fluence rate of (745$\pm$30)cm$^{-2}$d$^{-1}$. These neutrons that are produced by nuclear fission inside the reactor core, are reduced by a factor of $\sim$10$^{20}$ on their way to the CONUS shield. With a high-purity Ge detector without shield the $γ$-ray background was examined including highly thermal power correlated $^{16}$N decay products as well as $γ$-lines from neutron capture. Using the measured neutron spectrum as input, it was shown, with the help of Monte Carlo simulations, that the thermal power correlated field is successfully mitigated by the installed CONUS shield. The reactor-induced background contribution in the region of interest is exceeded by the expected signal by at least one order of magnitude assuming a realistic ionization quenching factor of 0.2.
△ Less
Submitted 31 December, 2019; v1 submitted 21 March, 2019;
originally announced March 2019.
-
Commissioning of the vacuum system of the KATRIN Main Spectrometer
Authors:
M. Arenz,
M. Babutzka,
M. Bahr,
J. P. Barrett,
S. Bauer,
M. Beck,
A. Beglarian,
J. Behrens,
T. Bergmann,
U. Besserer,
J. Blümer,
L. I. Bodine,
K. Bokeloh,
J. Bonn,
B. Bornschein,
L. Bornschein,
S. Büsch,
T. H. Burritt,
S. Chilingaryan,
T. J. Corona,
L. De Viveiros,
P. J. Doe,
O. Dragoun,
G. Drexlin,
S. Dyba
, et al. (125 additional authors not shown)
Abstract:
The KATRIN experiment will probe the neutrino mass by measuring the beta-electron energy spectrum near the endpoint of tritium beta-decay. An integral energy analysis will be performed by an electro-static spectrometer (Main Spectrometer), an ultra-high vacuum vessel with a length of 23.2 m, a volume of 1240 m^3, and a complex inner electrode system with about 120000 individual parts. The strong m…
▽ More
The KATRIN experiment will probe the neutrino mass by measuring the beta-electron energy spectrum near the endpoint of tritium beta-decay. An integral energy analysis will be performed by an electro-static spectrometer (Main Spectrometer), an ultra-high vacuum vessel with a length of 23.2 m, a volume of 1240 m^3, and a complex inner electrode system with about 120000 individual parts. The strong magnetic field that guides the beta-electrons is provided by super-conducting solenoids at both ends of the spectrometer. Its influence on turbo-molecular pumps and vacuum gauges had to be considered. A system consisting of 6 turbo-molecular pumps and 3 km of non-evaporable getter strips has been deployed and was tested during the commissioning of the spectrometer. In this paper the configuration, the commissioning with bake-out at 300°C, and the performance of this system are presented in detail. The vacuum system has to maintain a pressure in the 10^{-11} mbar range. It is demonstrated that the performance of the system is already close to these stringent functional requirements for the KATRIN experiment, which will start at the end of 2016.
△ Less
Submitted 3 March, 2016;
originally announced March 2016.
-
An angular-selective electron source for the KATRIN experiment
Authors:
M. Beck,
K. Bokeloh,
H. Hein,
S. Bauer,
H. Baumeister,
J. Bonn,
H. -W. Ortjohann,
B. Ostrick,
S. Rosendahl,
S. Streubel,
K. Valerius,
M. Zboril,
C. Weinheimer
Abstract:
The KATRIN experiment is going to search for the average mass of the electron antineutrino with a sensitivity of 0.2 eV/c2. It uses a retardation spectrometer of MAC-E filter type to accurately measure the shape of the electron spectrum at the endpoint of tritium beta decay. In order to achieve the planned sensitivity the transmission properties of the spectrometer have to be understood with high…
▽ More
The KATRIN experiment is going to search for the average mass of the electron antineutrino with a sensitivity of 0.2 eV/c2. It uses a retardation spectrometer of MAC-E filter type to accurately measure the shape of the electron spectrum at the endpoint of tritium beta decay. In order to achieve the planned sensitivity the transmission properties of the spectrometer have to be understood with high precision for all initial conditions. For this purpose an electron source has been developed that emits single electrons at adjustable total energy and adjustable emission angle. The emission is pointlike and can be moved across the full flux tube that is imaged onto the detector. Here, we demonstrate that this novel type of electron source can be used to investigate the transmission properties of a MAC-E filter in detail.
△ Less
Submitted 1 November, 2014;
originally announced November 2014.
-
Ellipsometry with polarisation analysis at cryogenic temperatures inside a vacuum chamber
Authors:
S. Bauer,
B. Grees,
D. Spitzer,
M. Beck,
R. Bottesch,
H. -W. Ortjohann,
B. Ostrick,
T. Schäfer,
H. H. Telle,
A. Wegmann,
M. Zbořil,
C. Weinheimer
Abstract:
In this paper we describe a new variant of null ellipsometry to determine thicknesses and optical properties of thin films on a substrate at cryogenic temperatures. In the PCSA arrangement of ellipsometry the polarizer and the compensator are placed before the substrate and the analyzer after it. Usually, the polarizer and the analyzer are rotated to find the intensity minimum searched for in null…
▽ More
In this paper we describe a new variant of null ellipsometry to determine thicknesses and optical properties of thin films on a substrate at cryogenic temperatures. In the PCSA arrangement of ellipsometry the polarizer and the compensator are placed before the substrate and the analyzer after it. Usually, the polarizer and the analyzer are rotated to find the intensity minimum searched for in null ellipsometry. In our variant we rotate the polarizer and the compensator instead, both being placed in the incoming beam before the substrate. Therefore the polarization analysis of the reflected beam can be realized by an analyzer at fixed orientation. We developed this method for investigations of thin cryogenic films inside a vacuum chamber, where the analyzer and detector had to be placed inside the cold shield at a temperature of T approx. 90K close to the substrate. All other optical components were installed at the incoming beam line outside the vacuum chamber, including all components which need to be rotated during the measurements. Our null ellipsometry variant has been tested with condensed krypton films on a highly oriented pyrolytic graphite substrate (HOPG) at a temperature of T approx. 25K. We show that it is possible to determine the indices of refraction of condensed krypton and of the HOPG substrate as well as thickness of krypton films with reasonable accuracy.
△ Less
Submitted 22 July, 2013;
originally announced July 2013.
-
Ultra-stable implanted 83Rb/83mKr electron sources for the energy scale monitoring in the KATRIN experiment
Authors:
M. Zbořil,
S. Bauer,
M. Beck,
J. Bonn,
O. Dragoun,
J. Jakůbek,
K. Johnston,
A. Kovalík,
E. W. Otten,
K. Schlösser,
M. Slezák,
A. Špalek,
T. Thümmler,
D. Vénos,
J. Žemlička,
C. Weinheimer
Abstract:
The KATRIN experiment aims at the direct model-independent determination of the average electron neutrino mass via the measurement of the endpoint region of the tritium beta decay spectrum. The electron spectrometer of the MAC-E filter type is used, requiring very high stability of the electric filtering potential. This work proves the feasibility of implanted 83Rb/83mKr calibration electron sourc…
▽ More
The KATRIN experiment aims at the direct model-independent determination of the average electron neutrino mass via the measurement of the endpoint region of the tritium beta decay spectrum. The electron spectrometer of the MAC-E filter type is used, requiring very high stability of the electric filtering potential. This work proves the feasibility of implanted 83Rb/83mKr calibration electron sources which will be utilised in the additional monitor spectrometer sharing the high voltage with the main spectrometer of KATRIN. The source employs conversion electrons of 83mKr which is continuously generated by 83Rb. The K-32 conversion line (kinetic energy of 17.8 keV, natural line width of 2.7 eV) is shown to fulfill the KATRIN requirement of the relative energy stability of +/-1.6 ppm/month. The sources will serve as a standard tool for continuous monitoring of KATRIN's energy scale stability with sub-ppm precision. They may also be used in other applications where the precise conversion lines can be separated from the low energy spectrum caused by the electron inelastic scattering in the substrate.
△ Less
Submitted 5 February, 2013; v1 submitted 20 December, 2012;
originally announced December 2012.
-
Prototype of an angular-selective photoelectron calibration source for the KATRIN experiment
Authors:
K. Valerius,
H. Hein,
H. Baumeister,
M. Beck,
K. Bokeloh,
J. Bonn,
F. Glück,
H. -W. Ortjohann,
B. Ostrick,
M. Zbořil,
Ch. Weinheimer
Abstract:
The method of direct neutrino mass determination based on the kinematics of tritium beta decay, which is adopted by the KATRIN experiment, makes use of a large, high-resolution electrostatic spectrometer with magnetic adiabatic collimation. In order to target a sensitivity on the neutrino mass of 0.2 eV/c^2, a detailed understanding of the electromagnetic properties of the electron spectrometer is…
▽ More
The method of direct neutrino mass determination based on the kinematics of tritium beta decay, which is adopted by the KATRIN experiment, makes use of a large, high-resolution electrostatic spectrometer with magnetic adiabatic collimation. In order to target a sensitivity on the neutrino mass of 0.2 eV/c^2, a detailed understanding of the electromagnetic properties of the electron spectrometer is essential, requiring comprehensive calibration measurements with dedicated electron sources. In this paper we report on a prototype of a photoelectron source providing a narrow energy spread and angular selectivity. Both are key properties for the characterisation of the spectrometer. The angular selectivity is achieved by applying non-parallel strong electric and magnetic fields: Directly after being created, photoelectrons are accelerated rapidly and non-adiabatically by a strong electric field before adiabatic magnetic guiding takes over.
△ Less
Submitted 29 November, 2010;
originally announced November 2010.
-
Effect of a sweeping conductive wire on electrons stored in the Penning trap between the KATRIN spectrometers
Authors:
M. Beck,
K. Valerius,
J. Bonn,
K. Essig,
F. Glück,
H. -W. Ortjohann,
B. Ostrick,
E. W. Otten,
Th. Thümmler,
M. Zbořil,
C. Weinheimer
Abstract:
The KATRIN experiment is going to search for the mass of the electron antineutrino down to 0.2 eV/c^2. In order to reach this sensitivity the background rate has to be understood and minimised to 0.01 counts per second. One of the background sources is the unavoidable Penning trap for electrons due to the combination of the electric and magnetic fields between the pre- and the main spectrometer at…
▽ More
The KATRIN experiment is going to search for the mass of the electron antineutrino down to 0.2 eV/c^2. In order to reach this sensitivity the background rate has to be understood and minimised to 0.01 counts per second. One of the background sources is the unavoidable Penning trap for electrons due to the combination of the electric and magnetic fields between the pre- and the main spectrometer at KATRIN. In this article we will show that by sweeping a conducting wire periodically through such a particle trap stored particles can be removed, an ongoing discharge in the trap can be stopped, and the count rate measured with a detector looking at the trap is reduced.
△ Less
Submitted 3 October, 2010; v1 submitted 18 September, 2009;
originally announced September 2009.
-
A UV LED-based fast-pulsed photoelectron source for time-of-flight studies
Authors:
K. Valerius,
M. Beck,
H. Arlinghaus,
J. Bonn,
V. M. Hannen,
H. Hein,
B. Ostrick,
S. Streubel,
Ch. Weinheimer,
M. Zboril
Abstract:
We report on spectroscopy and time-of-flight measurements using an 18 keV fast-pulsed photoelectron source of adjustable intensity, ranging from single photoelectrons per pulse to 5 photoelectrons per microsecond at pulse repetition rates of up to 10 kHz. Short pulses between 40 ns and 40 microseconds in length were produced by switching light emitting diodes with central output wavelengths of 2…
▽ More
We report on spectroscopy and time-of-flight measurements using an 18 keV fast-pulsed photoelectron source of adjustable intensity, ranging from single photoelectrons per pulse to 5 photoelectrons per microsecond at pulse repetition rates of up to 10 kHz. Short pulses between 40 ns and 40 microseconds in length were produced by switching light emitting diodes with central output wavelengths of 265 nm and 257 nm, in the deep ultraviolet (or UV-C) regime, at kHz frequencies. Such photoelectron sources can be useful calibration devices for testing the properties of high-resolution electrostatic spectrometers, like the ones used in current neutrino mass searches.
△ Less
Submitted 17 June, 2009; v1 submitted 13 February, 2009;
originally announced February 2009.
-
Long term stability of the energy of conversion electrons emitted from solid 83Rb/83mKr source
Authors:
D. Venos,
J. Kaspar,
M. Zboril,
O. Dragoun,
J. Bonn,
A. Kovalik,
O. Lebeda,
M. Rysavy,
K. Schlosser,
A. Spalek,
Ch. Weinheimer
Abstract:
The mono-energetic conversion electrons from the decay of 83mKr represent a unique tool for the energy calibration, energy scale monitoring and systematic studies of the tritium beta spectrum measurement in the neutrino mass experiment KATRIN. For this reason, the long term stability of energy of the 7.5 keV and 17.8 keV conversion electrons populated in the decay of solid 83Rb/83mKr vacuum evap…
▽ More
The mono-energetic conversion electrons from the decay of 83mKr represent a unique tool for the energy calibration, energy scale monitoring and systematic studies of the tritium beta spectrum measurement in the neutrino mass experiment KATRIN. For this reason, the long term stability of energy of the 7.5 keV and 17.8 keV conversion electrons populated in the decay of solid 83Rb/83mKr vacuum evaporated sources was examined by means of two electron spectrometers.
△ Less
Submitted 2 February, 2009;
originally announced February 2009.