Abstract
During laparoscopic surgery, the trocar insertion can injure arteries of the abdominal wall. Although these arteries are visible in a preoperative computed tomography [CT] with contrast medium, it is difficult for the surgeon to estimate their true intraoperative positions since the pneumoperitoneum dramatically stretches the abdominal wall. A navigation system showing the artery position would thus be very helpful for the surgeon. We present in this paper a method to simulate the position of the abdominal wall and its arteries after pneumoperitoneum. Our method requires a segmented preoperative CT image and an intraoperative surface reconstruction of the skin. The intraoperative skin surface allows us to compute a displacement field of the abdominal wall’s outer surface that we propagate to estimate the artery position.
Our simulation was evaluated using two sets of pig CT images, before and after pneumoperitoneum. Results show that our method provides an estimation of the abdominal wall and artery positions with an average error of respectively 2 mm and 6 mm which fits the clinical application constraint. In the near future, we will focus on viscera movement simulation after pneumoperitoneum using our abdominal wall shape prediction.
Access this chapter
Tax calculation will be finalised at checkout
Purchases are for personal use only
Preview
Unable to display preview. Download preview PDF.
Similar content being viewed by others
References
Sanchez-Margallo, F.M., et al.: Anatomical changes due to pneumoperitoneum analyzed by mri: an experimental study in pigs. Surg. Radiol. Anat. 33(5), 389–396 (2011)
Marescaux, J., et al.: Augmented-reality–assisted laparoscopic adrenalectomy. JAMA: the Journal of the American Medical Association 292(18), 2214 (2004)
Masamune, K., Sato, I., Liao, H., Dohi, T.: Non-metal Slice Image Overlay Display System Used Inside the Open Type MRI. In: Dohi, T., Sakuma, I., Liao, H. (eds.) MIAR 2008. LNCS, vol. 5128, pp. 385–392. Springer, Heidelberg (2008)
Nicolau, S.A., et al.: A cost effective simulator for education of ultrasound image interpretation and probe manipulation. Studies in Health Technology and Informatics 163, 403 (2011)
Saber, A.A., et al.: Safety zones for anterior abdominal wall entry during laparoscopy: a ct scan mapping of epigastric vessels. Annals of Surgery 239(2), 182 (2004)
Lam, A., et al.: Dealing with complications in laparoscopy. Best Practice & Research Clinical Obstetrics & Gynaecology 23(5), 631–646 (2009)
Geraci, G., et al.: Trocar-related abdominal wall bleeding in 200 patients after laparoscopic cholecistectomy: Personal experience. World Journal of Gastroenterology 12(44), 7165 (2006)
Kitasaka, T., Mori, K., Hayashi, Y., Suenaga, Y., Hashizume, M., Toriwaki, J.-i.: Virtual Pneumoperitoneum for Generating Virtual Laparoscopic Views Based on Volumetric Deformation. In: Barillot, C., Haynor, D.R., Hellier, P. (eds.) MICCAI 2004, Part II. LNCS, vol. 3217, pp. 559–567. Springer, Heidelberg (2004)
Shekhar, R., et al.: Live augmented reality: a new visualization method for laparoscopic surgery using continuous volumetric computed tomography. Surgical Endoscopy, 1–10 (2010)
Nakamoto, M., et al.: Recovery of respiratory motion and deformation of the liver using laparoscopic freehand 3d ultrasound system. Medical Image Analysis 11(5), 429–442 (2007)
Konishi, K., et al.: Augmented reality navigation system for endoscopic surgery based on three-dimensional ultrasound and computed tomography: Application to 20 clinical cases. International Congress Series, vol. 1281, pp. 537–542. Elsevier (2005)
Su, L.M., et al.: Augmented reality during robot-assisted laparoscopic partial nephrectomy: Toward real-time 3d-ct to stereoscopic video registration. Urology 73(4), 896–900 (2009)
Oda, M., et al.: Evaluation of deformation accuracy of a virtual pneumoperitoneum method based on clinical trials for patient-specific laparoscopic surgery simulator. In: Proceedings of SPIE, vol. 8316, p. 83160G (2012)
Hostettler, A., et al.: Real time simulation of organ motions induced by breathing: First evaluation on patient data. Biomedical Simulation, 9–18 (2006)
Nicolau, S.A., et al.: A structured light system to guide percutaneous punctures in interventional radiology. In: Proceedings of SPIE, vol. 7000, p. 700016 (2008)
Besl, P.J., et al.: A method for registration of 3-d shapes. IEEE Transactions on Pattern Analysis and Machine Intelligence 14(2), 239–256 (1992)
Author information
Authors and Affiliations
Editor information
Editors and Affiliations
Rights and permissions
Copyright information
© 2012 Springer-Verlag Berlin Heidelberg
About this paper
Cite this paper
Bano, J. et al. (2012). Simulation of the Abdominal Wall and Its Arteries after Pneumoperitoneum for Guidance of Port Positioning in Laparoscopic Surgery. In: Bebis, G., et al. Advances in Visual Computing. ISVC 2012. Lecture Notes in Computer Science, vol 7431. Springer, Berlin, Heidelberg. https://doi.org/10.1007/978-3-642-33179-4_1
Download citation
DOI: https://doi.org/10.1007/978-3-642-33179-4_1
Publisher Name: Springer, Berlin, Heidelberg
Print ISBN: 978-3-642-33178-7
Online ISBN: 978-3-642-33179-4
eBook Packages: Computer ScienceComputer Science (R0)