Abstract
A method is described by which bioelectrical spectroscopy could be combined with multifrequency electrical impedance tomography (EIT) enabling a Cole-Cole plot to be drawn for any selected pixel in the image for tissue characterization. To demonstrate the principles of the method, multifrequency EIT measurements were simulated with a numerical model. As an example, porcine myocardium, immersed in an electrolytic tank, was simulated for a frequency range of 20 kHz to 2 MHz. The method could have an important application in transplant surgery for screening organs for viability prior to transplant.