Main Memory Option Manual
Operating Documentation
CT HiSpeed Series & CT/e, CT ProSpeed AI/FI Series
GE MEDICAL SYSTEMS MAIN MEMORY OPTION
REV 4 2227162
a
CT HiSpeed Series & CT/e, CT ProSpeed AI/FI Series
GE MEDICAL SYSTEMS MAIN MEMORY OPTION
REV 4 2227162
b
CT HiSpeed Series & CT/e, CT ProSpeed AI/FI Series
GE MEDICAL SYSTEMS MAIN MEMORY OPTION
REV 4 2227162
In performing all electrical work on these products, GE will use its own specially trained field engineers. All of GE’s electrical
work on these products will comply with the requirements of the applicable electrical codes.
The purchaser of GE equipment shall only utilize qualified personnel (i.e., GE’s field engineers, personnel of third-party
service companies with equivalent training, or licensed electricians) to perform electrical servicing on the equipment.
c
CT HiSpeed Series & CT/e, CT ProSpeed AI/FI Series
GE MEDICAL SYSTEMS MAIN MEMORY OPTION
REV 4 2227162
DAMAGE IN TRANSPORTATION
All packages should be closely examined at time of delivery. If damage is apparent, have notation “damage in
shipment” written on all copies of the freight or express bill before delivery is accepted or “signed for” by a General
Electric representative or a hospital receiving agent. Whether noted or concealed, damage MUST be reported to the
carrier immediately upon discovery, or in any event, within 14 days after receipt, and the contents and containers held
for inspection by the carrier. A transportation company will not pay a claim for damage if an inspection is not requested
within this 14 day period.
Call Traffic and Transportation, Milwaukee, WI (414) 827–3449 / 8*285–3449 immediately after damage is found. At
this time be ready to supply name of carrier, delivery date, consignee name, freight or express bill number, item
damaged and extent of damage.
Complete instructions regarding claim procedure are found in Section “S” of the Policy & Procedure Bulletins.
GE personnel, please use the GEMS CQA Process to report all omissions, errors, and defects in this documentation.
Customers, please contact your GE Sales or Service represenatives.
CAUTION
Do not use the following devices near this equipment. Use of these devices near this equipment could cause
this equipment to malfunction.
Devices which intrinsically transmit radio waves such as; cellular phone, radio transceiver, mobile radio transmitter,
radio–controlled toy, etc.
Keep power to these devices turned off when near this equipment.
Medical staff in charge of this equipment is required to instruct technicians, patients and other people who may be
around this equipment to fully comply with the above regulation.
d
CT HiSpeed Series & CT/e, CT ProSpeed AI/FI Series
GE MEDICAL SYSTEMS MAIN MEMORY OPTION
REV 5 2227162
REVISION HISTORY
REV Date Primary Reason for Change
5 04/04/03 Updated Material List
A
CT HiSpeed Series & CT/e, CT ProSpeed AI/FI Series
GE MEDICAL SYSTEMS MAIN MEMORY OPTION
blank 2227162
B
CT HiSpeed Series & CT/e, CT ProSpeed AI/FI Series
GE MEDICAL SYSTEMS MAIN MEMORY OPTION
REV 4 2227162
TABLE OF CONTENTS
SECTION PAGE
i
CT HiSpeed Series & CT/e, CT ProSpeed AI/FI Series
GE MEDICAL SYSTEMS MAIN MEMORY OPTION
blank 2227162
ii
CT HiSpeed Series & CT/e, CT ProSpeed AI/FI Series
GE MEDICAL SYSTEMS MAIN MEMORY OPTION
REV 4 2227162
SECTION 1 – INTRODUCTION
This option provides DIMMs (memory) to increase a system main memory size.
FLOWCHART
The illustration below shows the main flow of the installation procedures of this option.
START
Preparation
DIMM Installation
Run Diagnostics
Scan Test
END
1–1
CT HiSpeed Series & CT/e, CT ProSpeed AI/FI Series
GE MEDICAL SYSTEMS MAIN MEMORY OPTION
blank 2227162
1–2
CT HiSpeed Series & CT/e, CT ProSpeed AI/FI Series
GE MEDICAL SYSTEMS MAIN MEMORY OPTION
REV 5 2227162
2236956–2
2236956
ITEM PART NO. DESCRIPTION
1 2235085 DIMM (O2 64MB x 2 Memory) ................................................ 4 4
2 2111577 Rating Plate ................................................................... 1
3 46–303268P1 Rating Plate (for GEHW) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
4 2227162 Instruction (This Manual) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 1
2–1
CT HiSpeed Series & CT/e, CT ProSpeed AI/FI Series
GE MEDICAL SYSTEMS MAIN MEMORY OPTION
blank 2227162
2–2
CT HiSpeed Series & CT/e, CT ProSpeed AI/FI Series
GE MEDICAL SYSTEMS MAIN MEMORY OPTION
REV 4 2227162
3-1 PREPARATION
1. Verify that the contents of this option kit are correct according to the Material List (see Section 2).
Front Cover
3. END
3–1
CT HiSpeed Series & CT/e, CT ProSpeed AI/FI Series
GE MEDICAL SYSTEMS MAIN MEMORY OPTION
REV 4 2227162
Special Equipment
D Tie Wraps
1. After turned the OC off, remove the following covers from the OC:
2. Loosen the two (2) screws (not necessary to remove) that fasten the lower part of the SCSI Bay (see Illustration
3–2).
3. Remove the OC Host Cover by unscrewing its six (6) screws (see Illustration 3–2).
4. Mark and disconnect the front connectors (signal and power cable connectors) of Host CPU Assy.
SCSI Bay
OC Host Cover
OC Host Cover Screws (x6)
5. END
3–2
CT HiSpeed Series & CT/e, CT ProSpeed AI/FI Series
GE MEDICAL SYSTEMS MAIN MEMORY OPTION
REV 4 2227162
NOTICE
Before touching any of the componens, attach the wrist strap to your wrist and to the metal
part of the chassis.
1. Disconnect the front connectors (signal and power cable connectors) from the System Module.
2. Push the ejection lever down to unseat the System Module from the backplane of the Host CPU assy.
Ejection Lever
Audio Assy
HDD
System Module
3–3
CT HiSpeed Series & CT/e, CT ProSpeed AI/FI Series
GE MEDICAL SYSTEMS MAIN MEMORY OPTION
REV 4 2227162
D For 512MB Option, remove all memories currently installed, then install the provided DIMMs into the slots
1~8.
b. Lower the DIMM, pushing it straight down into the socket gently but firmly.
You hear click as it is seated and the latches on each side of the module snap up.
5. Check the DIMMs visually to make sure they are installed correctly:
– The catches on the sides should be fitting snugly against the DIMM.
Memory Module
(DIMM)
3–4
CT HiSpeed Series & CT/e, CT ProSpeed AI/FI Series
GE MEDICAL SYSTEMS MAIN MEMORY OPTION
REV 4 2227162
Note
Stop for maintenance button can be clicked instead of pressing ESC key.
D Verify that the proper memory size is displayed: e.g. 512MB for 512MB option or 256MB for 128MB option.
D Take note of the messages underlined below to confirm that each memory test passed and completed
with no error.
3–5
CT HiSpeed Series & CT/e, CT ProSpeed AI/FI Series
GE MEDICAL SYSTEMS MAIN MEMORY OPTION
REV 4 2227162
2. Click on Reconfig button located on the upper right of the screen. The System configuration menu appears.
The system automatically reboots. Verify that the system starts up normally.
(It takes approximately 3 minutes.)
Note
If error occurs, the error log window appears. (You can confirm the problem by reviewing the error
log.) Click on Quit, the system automatically reboots again.
3–6
CT HiSpeed Series & CT/e, CT ProSpeed AI/FI Series
GE MEDICAL SYSTEMS MAIN MEMORY OPTION
REV 4 2227162
ÂÂ
ÂÂ
Â
ÂÂ
ÂÂ Â
ÂÂ
ÂÂ
ÂÂ
ÂÂ OC Rating Plate
Note
If other options have been installed in the system, several rating plates should have been attached.
In this case, attach this rating plate above the others.
4–1
CT HiSpeed Series & CT/e, CT ProSpeed AI/FI Series
GE MEDICAL SYSTEMS MAIN MEMORY OPTION
REV 4 2227162
2. Send the Product Locator Card to the appropriate address provided below.
4–2
GE Medical Systems: Telex 3797371
P.O. Box 414, Milwaukee, Wisconsin 53201 U.S.A.
(Asia, Pacific, Latin America, North America)