Abstract
Purpose
Revision hip arthroplasty using a modular tapered design gives the possibility for customising the prostheses to the individual anatomy intra-operatively. The success of this kind of surgery is still controversial due to the relative lack of medium- to long-term follow-up. Therefore we analysed the clinical and radiological outcome of the modular MRP-TITAN stem with diaphyseal fixation in revision hip surgery.
Methods
In this retrospective study we included 136 consecutive patients with MRP-TITAN stem implanted during revision hip arthroplasty. The average follow-up was 55 months. For clinical evaluation we used the Harris Hip Score and the Merle d’Aubigné and Postel score. The health-related quality of life was determined with the visual analogue pain scale.
Results
The surgeries were performed 109 months after primary total hip arthroplasty on average. The main indications for the MRP-TITAN revision stem were aseptic loosening, infection, and periprosthetic fracture. In the clinical outcome, patients achieved 75.1 points in the Harris Hip Score and 14.4 points in the Merle d’Aubigné and Postel Score. Mean level of persisting pain was 0.7 (VAS). The overall survival of the MRP stem in revision hip arthroplasty revealed 85.6 % survival at 9.75 years’ follow-up with a repeat revision rate of 6.8 %.
Conclusions
Performing revision hip arthroplasty using the MRP-TITAN stem revealed a good clinical outcome. There is a tendency for better results in comparison with the information given in literature for cementless modular revision stems including a lower rate in re-revisions.
Similar content being viewed by others
References
Fink B, Grossmann A, Schubring S, Schulz MS, Fuerst M (2009) Short-term results of hip revisions with a curved cementless modular stem in association with the surgical approach. Arch Orthop Trauma Surg 129:65–73. doi:10.1007/s00402-008-0617-7
Berry DJ (2002) Femoral revision: distal fixation with fluted, tapered grit-blasted stems. J Arthroplast 17:142–146
Mumme T, Muller-Rath R, Weisskopf M, Andereya S, Neuss M, Wirtz DC (2004) The cement-free modular revision prosthesis MRP-hip revision stem prosthesis in clinical follow-up. Z Orthop Grenzgeb 142:314–321. doi:10.1055/s-2004-822795
Gravius S, Randau T, Wirtz DC (2011) What can be done when hip prostheses fail?:new trends in revision endoprosthetics. Orthopade 40:1084–1094. doi:10.1007/s00132-011-1844-0
Wirtz DC, Gravius S, Ascherl R, Forst R, Noeth U, Maus UM, Zeiler G, Moritz CD (2014) Uncemented femoral revision arthroplasty using a modular tapered, fluted titanium stem. Acta Orthop 85:562–569. doi:10.3109/17453674.2014.958809
Mumme T, Muller-Rath R, Andereya S, Wirtz DC (2007) Uncemented femoral revision arthroplasty using the modular revision prosthesis MRP-TITAN revision stem. Oper Orthop Traumatol 19:56–77. doi:10.1007/s00064-007-1195-5
Kavanagh BF, Fitzgerald RH Jr (1985) Clinical and roentgenographic assessment of total hip arthroplasty. A new hip score. Clin Orthop Relat Res (193):133–140
Pak JH, Paprosky WG, Jablonsky WS, Lawrence JM (1993) Femoral strut allografts in cementless revision total hip arthroplasty. Clin Orthop Relat Res (295):172–178
Brooker AF, Bowerman JW, Robinson RA, Riley LH Jr (1973) Ectopic ossification following total hip replacement. Incidence and a method of classification. J Bone Joint Surg Am 55:1629–1632
Wagner H, Wagner M (1993) Femur revision prosthesis. Z Orthop Grenzgeb 131:574–577. doi:10.1055/s-2008-1040074
Bohm P, Bischel O (2001) Femoral revision with the Wagner SL revision stem: evaluation of one hundred and twenty-nine revisions followed for a mean of 4.8 years. J Bone Joint Surg Am 83-A:1023–1031
Randhawa K, Hossain FS, Smith B, Mauffrey C, Lawrence T (2009) A prospective study of hip revision surgery using the Exeter long-stem prosthesis: function, subsidence, and complications for 57 patients. J Orthop Traumatol 10:159–165. doi:10.1007/s10195-009-0068-0
Kim YH, Park JW, Kim JS (2013) Behaviour of the ultra-short anatomic cementless femoral stem in young and elderly patients. Int Orthop 37:2323–2330. doi:10.1007/s00264-013-2044-9
Schuh A, Werber S, Holzwarth U, Zeiler G (2004) Cementless modular hip revision arthroplasty using the MRP Titan Revision Stem: outcome of 79 hips after an average of 4 years’ follow-up. Arch Orthop Trauma Surg 124:306–309. doi:10.1007/s00402-004-0656-7
Wimmer MD, Randau TM, Deml MC, Ascherl R, Noth U, Forst R, Gravius N, Wirtz D, Gravius S (2013) Impaction grafting in the femur in cementless modular revision total hip arthroplasty: a descriptive outcome analysis of 243 cases with the MRP-TITAN revision implant. BMC Musculoskelet Disord 14:19. doi:10.1186/1471-2474-14-19
Gravius S, Mumme T, Andereya S, Maus U, Muller-Rath R, Wirtz DC (2007) The morse taper junction in modular revision hip replacement. Z Orthop Unfallchirurgie 145:157–160. doi:10.1055/s-2007-965182
Hernigou P, Queinnec S, Flouzat Lachaniette CH (2013) One hundred and fifty years of history of the Morse taper: from Stephen A. Morse in 1864 to complications related to modularity in hip arthroplasty. Int Orthop 37:2081–2088. doi:10.1007/s00264-013-1927-0
Wirtz DC, Heller KD, Holzwarth U, Siebert C, Pitto RP, Zeiler G, Blencke BA, Forst R (2000) A modular femoral implant for uncemented stem revision in THR. Int Orthop 24:134–138
Gutierrez Del Alamo J, Garcia-Cimbrelo E, Castellanos V, Gil-Garay E (2007) Radiographic bone regeneration and clinical outcome with the Wagner SL revision stem: a 5-year to 12-year follow-up study. J Arthroplast 22:515–524. doi:10.1016/j.arth.2006.04.029
van Diemen MP, Colen S, Dalemans AA, Stuyck J, Mulier M (2013) Two-stage revision of an infected total hip arthroplasty: a follow-up of 136 patients. Hip Int 23:445–450. doi:10.5301/hipint.5000049
Conflict of interest
The authors declare that they have no conflict of interest.
Author information
Authors and Affiliations
Corresponding author
Rights and permissions
About this article
Cite this article
Hoberg, M., Konrads, C., Engelien, J. et al. Outcome of a modular tapered uncemented titanium femoral stem in revision hip arthroplasty. International Orthopaedics (SICOT) 39, 1709–1713 (2015). https://doi.org/10.1007/s00264-015-2699-5
Received:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s00264-015-2699-5