Titel: Calibration of measuring instruments and standards for roughness metrology – Calibration of the horizontal measuring system of stylus instruments : Guideline DKD-R 4-2 Sheet 4
Autoren: Krüger-Sehm, Rolf, Bad Bevensen
Volk, Raimund, Jenoptik Industrial Metrology Germany GmbH, Villingen-Schwenningen
Beitragende: HostingInstitution: Physikalisch-Technische Bundesanstalt (PTB), ISNI: 0000 0001 2186 1887
Seiten:25
Sprache:en
DOI:10.7795/550.20240701
Version:03/2024
Art der Ressource: Text / Technical Requirement
Herausgeber: Physikalisch-Technische Bundesanstalt (PTB)
Rechte: https://creativecommons.org/licenses/by-nc-nd/3.0/de/
CC by-nc-nd 3.0
Beziehungen: IsVariantFormOf: DOI 10.7795/550.20240411
Daten: Verfügbar: 2024-07-03
Erstellt: 2024-03
Datei: Datei herunterladen (application/pdf) 1.32 MB (1381837 Bytes)
MD5 Prüfsumme: 41c89a5009d944cb49046a6d96bad67e
SHA256 Prüfsumme: 2b66f089513722038f3bc21e274f7b02df9ce3967fcb6ce1cfa2da27cd84d6c9
Schlagworte DKD Guideline ; roughness measurement ; stylus instrument ; ISO 3274 ; horizontal axis ; metrological traceability
Zusammenfassung: This guideline describes how to establish traceability for the scaling of the horizontal axis of stylus instruments to the national length standard in accordance with ISO 3274. The metrological traceability is based on the determination of the value of a horizontal surface parameter (characteristic quantity) such as RSmon a calibrated reference standard. The unit of length is passed on by alternately using a standard for calibrating a device, and a calibrated device for calibrating a standard. The scope of this guideline applies to the second and to the fourth step in the dissemination chain (described in the guideline sheet in Appendix A and Appendix B, respectively).
Zitierform: Guideline DKD-R 4-2 Calibration of measuring instruments and standards for roughness metrology, Sheet 4 Calibration of the horizontal measuring system of stylus instruments, Edition 03/2024, Revision 0, Physikalisch-Technische Bundesanstalt, Braunschweig and Berlin. DOI: 10.7795/550.20240701