Development of a Symmetric EPI Framework for Clinical Translation of Rapid Dynamic Hyperpolarized 13C Imaging

Jeremy W Gordon 1, Daniel B Vigneron 1, Peder EZ Larson 1
PMCID: PMC4992668  NIHMSID: NIHMS748229  PMID: 26898849

Abstract

Purpose

To develop symmetric echo-planar imaging (EPI) and a reference scan framework for hyperpolarized 13C metabolic imaging.

Methods

Symmetric, ramp-sampled EPI with partial Fourier reconstruction was implemented on a 3T scanner. The framework for acquiring a reference scan on the 1H channel and applied to 13C data was described and validated in both phantoms and in vivo metabolism of [1-13C]pyruvate.

Results

Ramp-sampled, symmetric EPI provided a substantial increase in the SNR of the phantom experiments. The reference scan acquired on the 1H channel yielded 13C phantom images that varied in mean signal intensity <2%, as compared to 13C images reconstructed with a reference scan directly measured on the 13C channel. The structural similarity index and dynamic time course from in vivo 13C data further support the application of a 1H reference scan to 13C data to mitigate Nyquist ghost artifacts.

Conclusion

Ramp-sampled, symmetric EPI with spectral-spatial excitation of a single metabolite provides a fast, robust and clinically efficacious approach to acquire hyperpolarized 13C dynamic molecular imaging data. The gains of this efficient sampling, combined with partial Fourier methods, enables large matrix sizes required for human studies.

Introduction

Advances in hyperpolarization (HP) have enabled real-time molecular imaging that was heretofore impossible with magnetic resonance (1). Using the dynamic nuclear polarization (DNP) technique, spins are exogenously polarized to four orders of magnitude greater than Boltzmann equilibrium. Metabolically active substrates, such as [1-13C]pyruvate (2), are transported into cells and undergo enzymatic conversion into downstream metabolites on the timescale of an HP experiment (~1-2 minutes). This technique is being translated into the clinic, primarily for cancer and cardiac metabolism studies. A Phase 1 clinical trial demonstrated both safety and the feasibility of hyperpolarized [1-13C]pyruvateMR with up-regulated conversion to [1-13C]lactate conversion observed in biopsy-proven prostate cancer patients (3).

This first clinical trial used a 13C spectroscopic data acquisition method, which even with an echo-planar spectroscopic readout, required an acquisition time of 8-12 seconds for 3D volumetric coverage of the prostate with a spatial resolution of 0.34cm3.This spectroscopic data demonstrated increased HP [1-13C]lactate to [1-13C]pyruvate ratios in the prostate cancer patients, but the acquisition method had the drawbacks of relatively long temporal resolution, preventing accurate dynamic measurements of metabolism, and the complex EPSI reconstruction was performed offline, not on the clinical scanner.A different approach for rapid volumetric HP 13C MRI acquisition is selective excitation of a single resonance followed by a single shot flyback EPI readout (4). This technique was developed and applied to study the dynamics of [1-13C]lactate in preclinical cancer models following injection of [1-13C]pyruvate with a 3.5s temporal resolution over a 100s time window (4). Although this study showed significant differences in [1-13C]lactate dynamics between early and late stage transgenic prostate cancers, without the detection of [1-13C]pyruvate signals the measurement of conversion rates could not be made. Utilizing the chemical shift differences between the lactate and pyruvate resonances, acquisitions of both compounds can be acquired simultaneously (5), but suffer from FOV restrictions and sub-optimal SNR.

A major challenge facing these HP acquisition techniques is the limited imaging window, an outcome of the inexorable decay to the Boltzmann (or thermal) equilibrium polarization. Rapid and RF efficient imaging sequences are therefore of paramount importance to maximize the signal-to-noise ratio (SNR) and resolution in 13C experiments. Using a singleband spectral-spatial RF pulse (4,6), individual metabolites can be excited and rapidly imaged with a single excitation. Due to the slow transverse decay of 13C-labeled metabolites (7), a low bandwidth/long duration readout can be used to maximize SNR efficiency. An echo-planar imaging (EPI) readout trajectory in particular is well-suited for hyperpolarized imaging because of its rapid, efficient sampling and relative robustness to off-resonance artifacts. Flyback EPI trajectories have been used (5,8) for rapid metabolic imaging due to their insensitivity to timing delays and ease of implementation, but this approach suffers from a reduced SNR efficiency because of the lack of encoding during the flyback gradient. The lower 13C gyromagnetic ratio also results in reduced scan efficiency and longer echo times (TE) that exacerbate EPI-related image artifacts. Translating this technique to clinical 13C imaging requires further improvements in scan efficiency in order to improve SNR and enable rapid and robust metabolic imaging in vivo.

Symmetric EPI is an appealing alternative to flyback EPI. By encoding k-space on both the positive and negative gradient lobes, the echo-spacing can be substantially reduced, resulting in improved SNR and less sensitivity to off-resonance. The SNR efficiency is also increased with symmetric EPI due to a higher sampling duty-cycle, and can be further increased by using ramp-sampling. The downside of this approach is that symmetric EPI is prone to Nyquist ghosting artifacts because of delays and inconsistencies between even and odd lines of k-space, which primarily arise from eddy currents and concomitant magnetic fields (9). These errors manifest as an offset between even and odd lines of k-space, resulting in misalignment between the even and odd echoes and resulting in image ghosts that are shifted by half the FOV if left uncorrected (10). These timing delays and inconsistencies can be estimated with a reference scan (11,12), and, by applying a line-by-line phase correction in x-ky space, Nyquist ghost artifacts can be minimized. However, a reference scan will be difficult to directly acquire from the hyperpolarized substrate due to the transient magnetization and the narrow window in which to image metabolism.

The goal of this project was to develop and test a new specialized HP 13C EPI acquisition approach with high SNR efficiency and temporal resolution that could be readily translated into future human studies. To that end, we developed and implemented a symmetric, ramp-sampled, partial Fourier EPI sequence for efficient HP 13C imaging. We also developed a framework for acquiring a reference scan from 1H data for application to hyperpolarized 13C data on a clinical 3T MR scanner. Our approach was designed to be robust and SNR-efficient, have reduced EPI off-resonance artifacts, and to eliminate EPI ghosting artifacts, making it well-suited for clinical metabolic imaging studies with hyperpolarized 13C agents.

Methods

All experiments were performed on a clinical 3T scanner (MR 750, GE Healthcare, WI) with a maximum gradient strength of 5 G/cm and a maximum slew-rate of 20 G/cm/ms. A ramp-sampled, symmetric EPI readout with singleband spectral-spatial (SPSP) excitation and partial Fourier reconstruction was developed for use on the clinical GE platform. A reference scan, acquired prior to imaging by disabling the phase-encoding gradients, was used to correct for phase errors that lead to Nyquist ghosts. As the sequence was built off of a product GE sequence, it leveraged the ability to use the reference scan for online reconstruction of 13C metabolite maps, and retained all of the requisite safety and performance features.

A custom built dual-tune 1H/13C volume coil (13) was used for both the thermal phantom and hyperpolarized 13C experiments. For in vivo experiments, an enriched 8M 13C urea phantom was used for center frequency and B1 calibration. The SPSP RF pulse used in all of the EPI acquisitions was designed to excite [1-13C]pyruvate and metabolites at 3T, with a passband FWHM of 120Hz and a stopband of 600Hz. The RF pulse was designed in Matlab (R2014b) using the SPSP RF pulse design toolbox (14). Further information and downloads can be found at http://rsl.stanford.edu/research/software.html or https://github.com/agentmess/hyperpolarized-mri-toolbox.

The SNR efficiency is a function of the duty cycle and k-space trajectory, and will increase from a flyback to a symmetric EPI readout. This gain in SNR efficiency, which is equivalent to the product of the duty cycle and trajectory efficiency, can be defined as (15,16):

SNReff,duty=TreadoutTgradient [1]
SNReff,unifornity=AkkD(k)dkk1D(k)dk [2]

For flyback EPI, Tgradient is the duration of the trapezoidal readout gradient plus the flyback gradient. For symmetric EPI, Tgradient is the duration of the trapezoidal readout gradient. Treadout is the time the analog-to-digital converter is acquiring data. For non ramp-sampled EPI, Treadout is equal to the plateau of the readout gradient, while for ramp-sampled EPI, Treadout is equal to Tgradient minus the duration of the phase encode blip. Ak is the area of the sampled k-space, and Dk is the local k-space sampling density. For partial Fourier, the theoretical SNR efficiency is reduced by 1.15 to account for the reduction in acquisition time for 75% k-space coverage. To quantify the SNR advantages of symmetric EPI, a 13C enriched phantom was used (44mm outer diameter (OD)), consisting of four cylindrical tubes with either 1M 13C formic acid (14.8mm OD), 1M [1-13C]lactate (11.9mm OD), 1M [1-13C]alanine (10.2mm OD) or 1M 13C bicarbonate (19.6mm OD) to simulate the expected number of metabolites and chemical shift in a HP 13C experiment. The RF pulse was centered on bicarbonate, and data were acquired with either a symmetric or flyback echo-planar trajectory, with a 1s TR, a FOV of 96 × 96mm, a matrix size of 32 × 32, and 60 averages, yielding a total 1 minute scan time. For the symmetric readout (Fig. 1) data were either fully encoded or acquired with a ramp-sampled, symmetric readout with partial Fourier reconstruction (75% k-space coverage in the ky direction; Table 1). A reference scan was obtained directly on the 13C channel by disabling the phase-encoding gradients and was acquired prior to imaging using 60 averages. Flyback EPI was reconstructed with a 2D Fourier Transform, while for ramp-sampled symmetric EPI the reference scan phase correction was applied (17), then data were interpolated to a Cartesian grid and 2D Fourier Transformed to image-space using the Orchestra reconstruction toolbox (GE Healthcare, WI, USA). The SNR was measured by calculating the mean signal within the phantom and then dividing by the standard deviation of the signal from a noise ROI. No changes to the Orchestra EPI routines were required for our 13C EPI application, and the routines are fully automatic once the location of the reference scan and data are specified.

Figure 1.

Figure 1

Diagram of the symmetric EPI pulse sequence used in this work. The sequence consists of a singleband SPSP RF excitation, followed by a symmetric EPI readout. Phase errors due to timing delays are removed by a separate reference scan on the 1H channel, accomplished by disabling the phase-encoding gradients.

Table 1.

SNR and scan parameter comparison between EPI modes for a 96 x 96mm FOV and a 32 × 32 matrix. The improvement in measured SNR for the symmetric EPI modes is due to a higher duty cycle as well as shorter echo time.

TE Echo
Spacing
Duty Cycle Trajectory
Efficiency
Relative SNR
(Measured)
Flyback EPI 57.9 ms 2.24 ms 0.534 1.0 0.50 ± 0.03
Symmetric EPI 45.2 ms 1.74 ms 0.768 1.0 0.91 ± 0.07
Symmetric EPI
with ramp
sampling and
partial-Fourier
26.4 ms 1.14 ms 0.92 0.879 1.0 ± 0.08

Unlike flyback EPI, symmetric EPI requires a reference scan to correct for timing delays and inconsistencies between even and odd lines of k-space. This can potentially be acquired on either the 1H channel or directly on the 13C channel.Due to the lack of endogenous 13C, using the 13C channel requires either an external 13C enriched syringe or expending some hyperpolarized magnetization. To determine if a 1H reference scan could provide a robust correction for 13C images, reference scans were acquired on either the 1H channel or on the 13C channel from a natural abundance ethylene glycol phantom (outer diameter of 54mm, inner diameter of 24 mm) using the same trajectory (designed for a 13C acquisition). The singleband SPSP pulse described above was used to avoid excitation of the J-coupled 13C resonances. The matrix size and FOV were 32 × 32 and 72 × 72 mm, respectively, with partial Fourier (75% k-space coverage) in the phase-encode direction and ramp-sampling along the readout (0.492ms ramp duration, 0.380ms plateau duration, 1.364ms echo-spacing). A single 50 mm thick slice was excited, with 360 averages, a 0.5 s TR, and a 21.9 ms TE, with a total readout duration of 32.7 ms. Reference scans were applied to the 13C data retroactively using the Orchestra Toolbox, and image quality was compared by calculating the fractional difference between the phantom data reconstructed with the two different reference scans.

To explore the utility of the 1H reference scan for in vivo HP 13C dynamic imaging, metabolites maps of pyruvate and lactate were acquired in the study of renal metabolism in a healthy rat. Metabolites were excited with a singleband SPSP RF pulse, with independent flip-angles designed to achieve the best estimates of perfusion and metabolic conversion rate parameters in a physiological model (18). A total of 25 timeframes for each metabolite were acquired over 50s, for a temporal resolution of 2s, with a FOV of 96 × 96mm and a matrix size of 32 × 32. Partial Fourier (75% k-space coverage) was used in the phase-encode direction along with ramp-sampling (0.452ms ramp duration, 0.240ms plateau duration, 1.144ms echo-spacing) to decrease the echo-time, yielding a TE of 20 ms with a 100ms TR per metabolite. Data acquisition started immediately prior to injection covering a 20mm axial slice centered on the kidneys. A reference scan was acquired on the 1H channel using the 13C trajectory prior to injection of the HP substrate. For comparison, a separate reference scan was directly acquired on the 13C channel from the remaining hyperpolarized substrate magnetization following the final image. The efficacy of the 1H reference scan was determined by calculating the structural similarity (SSIM) index (19) of [1-13C]pyruvate for all time frames, and by comparing the time course between data reconstructed with either the 1H or 13C reference scan. For the SSIM index, the 13C images reconstructed with the 13C reference scan were considered the ground truth. SSIM is a measure of image quality and signal fidelity between two images, with an SSIM of 1 implying complete agreement between the two datasets. SSIM was chosen as the metric of comparison because it has been shown to be more sensitive to image shifts and overall changes in signal than mean square error (19,20).

Results

The results from the phantom study are shown in Table 1 and Supporting Figure 1. The substantial SNR increase between flyback and symmetric EPI is due to the higher duty cycle for symmetric EPI, which encodes k-space on both the positive and negative readout lobes and reduces the TE. Further gains were observed when using ramp-sampling and a partial Fourier acquisition, as this further reduces the echo-time and increases the readout duty cycle, mitigating signal loss due to T2* decay.

The 1H reference scan provides a potential alternative method that is more robust and has higher SNR than direct acquisition on the 13C channel. The reference scan in both cases was acquired using the 13C trajectory, resulting in the same echo-spacing and timing delays for both approaches. The 13C phantom images reconstructed without a reference scan show substantial ghosting (Fig. 2A). Qualitatively, the lack of Nyquist ghosting in the phantom images reconstructed with either the 1H or 13C reference scan (Fig. 2B,C) demonstrates the effectiveness of this technique. Quantitatively, the mean signal difference between both reconstruction approaches was <2% for all phantom voxels, further indicating that a 1H reference scan was sufficient to correct for Nyquist ghosting in 13C images.

Figure 2.

Figure 2

The utility of a 1H reference scan for 13C EPI data. The uncorrected 13C EPI image (A) exhibits strong Nyquist ghost artifacts in the absence of a reference scan. In contrast, 13C phantom images reconstructed with either the 1H (B) or 13C (C) reference scan exhibit minimal ghosting, with a mean signal difference <2% between both reference scans for all phantom voxels. Using a 1H reference scan is advantageous because it requires no external 13C reference scan nor uses any hyperpolarized magnetization.

This approach was further extended to the in vivo study of [1-13C]pyruvate renal metabolism in a healthy rat. Images reconstructed with either a 13C reference scan obtained from the residual hyperpolarization (Fig. 3A) or a 1H reference scan obtained prior to 13C imaging (Fig. 3B) are free of Nyquist ghosts. The SSIM index (Fig. 3D) was measured to be greater than 0.91 for all timeframes using the 1H reference scan, whereas the absence of a reference scan leads to Nyquist ghosting (Fig. 3C) and substantially lower structural similarity for all timeframes. The dynamic time course (Fig. 3E) was not significantly different (t-test, p = 0.98), further supporting the utility of a 1H reference scan.

Figure 3.

Figure 3

In vivo validation of 1H reference scan for 13C EPI data. Representative images reconstructed with either the 13C reference scan (A) or 1H reference scan (B) are free of Nyquist ghosts, while the lack of a reference scan (C) clearly results in artifacts (arrows). The structural similarity index (SSIM; D) for pyruvate images reconstructed without a reference scan is substantially lower than data reconstructed using the 1H reference scan. The ground truth was considered to be images reconstructed with the 13C reference scan. The time course of pyruvate and lactate (E) is nearly identical for data reconstructed using either the 1H or 13C reference scan (< 0.5% difference).

The in vivo dynamic data from a separate experiment can be seen in Fig. 4. Both pyruvate and lactate metabolite maps exhibit high SNR and are free of Nyquist ghost artifacts. The variable flip-angle schedule attempts to optimize the rate parameter estimates in vivo, yielding lactate images that provide nearly constant SNR throughout the time course. The absence of ghosting throughout the acquisition indicates that gradient heating and infidelity was not an issue, and a single reference scan was sufficient for phase correction for all time points. The higher spatial resolution of EPI (3mm in-plane) also enabled the visualization of differential metabolism between the renal cortex and medulla.

Figure 4.

Figure 4

Dynamic images of renal metabolism acquired with symmetric EPI. Both pyruvate and lactate metabolites images were free of Nyquist ghost artifacts and exhibit high SNR. Note the detail of differential metabolism between the renal cortex and medulla in the early pyruvate and lactate images that was enabled by the higher resolution EPI acquisition. Images have been zero-filled from 32×32 to 128×128 for display purposes.

Discussion

In this study, we implemented and tested a ramp-sampled, symmetric, partial Fourier EPI sequence with spectral-spatial RF excitation for hyperpolarized 13C imaging. The benefits of a symmetric readout were validated, and the feasibility of obtaining a suitable reference scan and artifact-free metabolite maps in vivo were demonstrated. In particular, ramp-sampling and partial Fourier further reduced the echo-spacing and TE for symmetric EPI, making it more robust to chemical shift artifacts in the slow phase encoding direction. In principle ramp-sampling and partial Fourier can result in an SNR penalty because of the reduction in total acquisition time and trajectory efficiency, but in practice this reduction is typically offset by the increase in SNR from the shorter TE and reduced echo-spacing. This acceleration will be crucial for the translation to clinical imaging, as the larger patient FOVs will necessitate a concomitant increase in matrix size to maintain adequate spatial resolution.

Estimating the timing delays between even and odd lines of k-space is crucial to minimizing ghosting artifacts in a symmetric EPI acquisition. We explored the reference scan correction using the 1H channel, an external 13C urea syringe, or the remaining magnetization from the hyperpolarized substrate. The external 13C syringe works well for single, thick slice phantom studies but is problematic for thinner slices due to poor SNR or for multi-slice experiments when the imaging volume is larger than the phantom. While the reference scan using the residual hyperpolarization can also produce a suitable reference scan, it worked best in conjunction with a single-slice axial technique. This is an outcome of the experimental setup, since in-flowing spins provide fresh magnetization for a reference scan following a VFA scheme that ends at a 90-degree flip. For coronal slices, or for 3D imaging, this approach will likely prove difficult because of the lack of residual magnetization.

Instead, acquiring the reference scan from 1H data is an appealing alternative. Complementary information from the 1H channel has been previously shown to aid in 13C reconstruction. For instance, a field map can be readily obtained on the 1H channel and applied to 13C data by accounting for the difference in gyromagnetic ratio (21), and the k-space trajectory can be measured using Duyn’s method (22) on the 1H channel and applied to 13C data in a similar manner. Similarly, we have shown that a suitable reference scan can be obtained in an analogous manner. This approach provided a more robust, higher SNR method to calculate the phase coefficients and correct for timing delays. The similarity in the computed coefficients, as well as the structural similarity index between images reconstructed with the 1H or 13C reference scan, implies that a 1H reference scan can be acquired prior to and applied on hyperpolarized 13C imaging to generate images that are free of Nyquist ghost artifacts. This technique is likely to be more robust than external 13C syringes due to improved SNR and will be especially useful in clinical imaging where space considerations and patient comfort must be considered.

Other methods of accounting for timing delays in symmetric EPI have limitations. Doubly sampled EPI (23) can yield artifact free images by acquiring each line of k-space twice and separately reconstructing even and odd echoes. However this will substantially increase the echo-spacing and echo-time, resulting in signal loss and further filtering due to T2* decay. Timing delays can also be estimated by acquiring two images, with the second shifted by one echo such that the even lines of k-space in the first image line up with the odd lines of k-space in the second image (24). This method effectively doubles the scan time and RF deposition, rendering it largely ineffective for hyperpolarized imaging. Navigator echoes are an alternative approach (25) that have also been shown to be effective in minimizing Nyquist ghosts. While it provides an updated reference scan for each timeframe, it does results in an increase in the TE. Calibrationless reconstruction techniques, such as minimizing the entropy in image-space (26), are an appealing alternative to a reference scan to mitigate Nyquist ghosting and will be explored in subsequent research.

Both spiral and Cartesian EPI are well-suited for single-shot imaging of hyperpolarized substrates, as both utilize high duty cycle and efficient trajectories (27-29). While a spiral acquisition has the benefit of a shorter echo-time and high SNR efficiency, it is more sensitive to off-resonance artifacts and gradient infidelities, but is robust to flow effects and pulsatile motion (30,31). Hyperpolarized pyruvate studies using spiral trajectories have relied on auto-focusing reconstruction methods to correct for off-resonance artifacts (8). EPI is more robust in the presence of off-resonance, resulting in either a benign shift in the PE direction for chemical shift or object distortion for an inhomogeneous B0 field. Partial Fourier and parallel imaging techniques are also readily amenable to EPI trajectories and can be used to accelerate imaging (32), further reduce RF deposition, and reduce the TE for larger FOVs.

Conclusion

This ramp-sampled, symmetric EPI approach with spectral-spatial excitation of a single metabolite provided a fast, robust and clinically efficacious way to acquire hyperpolarized 13C dynamic molecular imaging data. The substantial?SNR increase enables improved temporal and spatial localization and?increased scan coverage. The gains of this efficient sampling, combined with?partial Fourier methods, will be crucial for large matrix sizes required for?human-sized FOVs. This pulse sequence is well suited for clinical dynamic imaging of hyperpolarized substrates as it leverages well-established EPI methods to reconstruct 13C metabolite maps on a commercial scanner in real time, and can be easily incorporated into existing diagnostic workflows.

Supplementary Material

Supp Fig S1

Acknowledgements

We would like to thank Dr. Robert Bok for help with animal handling, Ralph Hurd and Wenwen Jiang for helpful discussions, and John Maidens for providing the dynamic flip angle designs.This work was supported by NIH grants R01 EB017449, R01 EB016741, and P41 EB013598.

References

  • 1.Ardenkjær-Larsen JH, Fridlund B, Gram A, Hansson G, Hansson L, Lerche MH, Servin R, Thaning M, Golman K. Increase in signal-to-noise ratio of > 10,000 times in liquid-state NMR. Proc Natl Acad Sci USA. 2003;100(18):10158–10163. doi: 10.1073/pnas.1733835100. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 2.Golman K, Olsson LE, Axelsson O, Mansson S, Karlsson M, Petersson JS. Molecular imaging using hyperpolarized 13C. Br J Radiol. 2003;76(suppl_2):S118–127. doi: 10.1259/bjr/26631666. [DOI] [PubMed] [Google Scholar]
  • 3.Nelson SJ, Kurhanewicz J, Vigneron DB, Larson PEZ, Harzstark AL, Ferrone M, van Criekinge M, Chang JW, Bok R, Park I, Reed G, Carvajal L, Small EJ, Munster P, Weinberg VK, Ardenkjaer-Larsen JH, Chen AP, Hurd RE, Odegardstuen L-I, Robb FJ, Tropp J, Murray JA. Metabolic Imaging of Patients with Prostate Cancer Using Hyperpolarized [1-13C]Pyruvate. Science Translational Medicine. 2013;5(198):198ra108. doi: 10.1126/scitranslmed.3006070. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 4.Lupo JM, Chen AP, Zierhut ML, Bok RA, Cunningham CH, Kurhanewicz J, Vigneron DB, Nelson SJ. Analysis of hyperpolarized dynamic 13C lactate imaging in a transgenic mouse model of prostate cancer. Magnetic Resonance Imaging. 2010;28(2):153–162. doi: 10.1016/j.mri.2009.07.007. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 5.Reed GD, Larson PEZ, Morze Cv, Bok R, Lustig M, Kerr AB, Pauly JM, Kurhanewicz J, Vigneron DB. A method for simultaneous echo planar imaging of hyperpolarized 13C pyruvate and 13C lactate. Journal of Magnetic Resonance. 2012;217(0):41–47. doi: 10.1016/j.jmr.2012.02.008. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 6.Cunningham CH, Chen AP, Lustig M, Hargreaves BA, Lupo J, Xu D, Kurhanewicz J, Hurd RE, Pauly JM, Nelson SJ, Vigneron DB. Pulse sequence for dynamic volumetric imaging of hyperpolarized metabolic products. Journal of Magnetic Resonance. 2008;193(1):139–146. doi: 10.1016/j.jmr.2008.03.012. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 7.Cunningham CH, Chen AP, Albers MJ, Kurhanewicz J, Hurd RE, Yen Y-F, Pauly JM, Nelson SJ, Vigneron DB. Double spin-echo sequence for rapid spectroscopic imaging of hyperpolarized 13C. Journal of Magnetic Resonance. 2007;187(2):357–362. doi: 10.1016/j.jmr.2007.05.014. [DOI] [PubMed] [Google Scholar]
  • 8.Lau AZ, Chen AP, Ghugre NR, Ramanan V, Lam WW, Connelly KA, Wright GA, Cunningham CH. Rapid multislice imaging of hyperpolarized 13C pyruvate and bicarbonate in the heart. Magnetic Resonance in Medicine. 2010;64(5):1323–1331. doi: 10.1002/mrm.22525. [DOI] [PubMed] [Google Scholar]
  • 9.Du YP, Joe Zhou X, Bernstein MA. Correction of concomitant magnetic field-induced image artifacts in nonaxial echo-planar imaging. Magnetic Resonance in Medicine. 2002;48(3):509–515. doi: 10.1002/mrm.10249. [DOI] [PubMed] [Google Scholar]
  • 10.Le Bihan D, Poupon C, Amadon A, Lethimonnier F. Artifacts and pitfalls in diffusion MRI. Journal of Magnetic Resonance Imaging. 2006;24(3):478–488. doi: 10.1002/jmri.20683. [DOI] [PubMed] [Google Scholar]
  • 11.Bruder H, Fischer H, Reinfelder HE, Schmitt F. Image reconstruction for echo planar imaging with nonequidistant k-space sampling. Magnetic Resonance in Medicine. 1992;23(2):311–323. doi: 10.1002/mrm.1910230211. [DOI] [PubMed] [Google Scholar]
  • 12.Ahn CB, Kim JH, Cho ZH. High-Speed Spiral-Scan Echo Planar NMR Imaging-I. Medical Imaging, IEEE Transactions on. 1986;5(1):2–7. doi: 10.1109/TMI.1986.4307732. [DOI] [PubMed] [Google Scholar]
  • 13.Derby K, Tropp J, Hawryszko C. Design and evaluation of a novel dual-tuned resonator for spectroscopic imaging. Journal of Magnetic Resonance (1969) 1990;86(3):645–651. [Google Scholar]
  • 14.Larson PEZ, Kerr AB, Chen AP, Lustig MS, Zierhut ML, Hu S, Cunningham CH, Pauly JM, Kurhanewicz J, Vigneron DB. Multiband excitation pulses for hyperpolarized 13C dynamic chemical-shift imaging. Journal of Magnetic Resonance. 2008;194(1):121–127. doi: 10.1016/j.jmr.2008.06.010. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 15.Tsai C-M, Nishimura DG. Reduced aliasing artifacts using variable-density k-space sampling trajectories. Magnetic Resonance in Medicine. 2000;43(3):452–458. doi: 10.1002/(sici)1522-2594(200003)43:3<452::aid-mrm18>3.0.co;2-b. [DOI] [PubMed] [Google Scholar]
  • 16.Jiang W, Lustig M, Larson PEZ. Concentric rings K-space trajectory for hyperpolarized 13C MR spectroscopic imaging. Magnetic Resonance in Medicine. 2014 doi: 10.1002/mrm.25577. DOI: 10.1002/mrm.25577. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 17.Hinks R, Mock B, Collick B, Frigo F, Shubhachint T. Method and system for image artifact reduction using nearest-neighbor phase correction for echo planar imaging. Google Patents. 2006 [Google Scholar]
  • 18.Maidens J, Larson PEZ, Arcak M. Optimal Experiment Design for Physiological Parameter Estimation Using Hyperpolarized Carbon-13 Magnetic Resonance Imaging; Proceedings of the American Control Conference; Chicago, IL. 2015. FrC11.1. [Google Scholar]
  • 19.Zhou W, Bovik AC, Sheikh HR, Simoncelli EP. Image quality assessment: from error visibility to structural similarity. Image Processing, IEEE Transactions on. 2004;13(4):600–612. doi: 10.1109/tip.2003.819861. [DOI] [PubMed] [Google Scholar]
  • 20.Zhou W, Bovik AC. Mean squared error: Love it or leave it? A new look at Signal Fidelity Measures. Signal Processing Magazine, IEEE. 2009;26(1):98–117. [Google Scholar]
  • 21.Gordon JW, Niles DJ, Fain SB, Johnson KM. Joint spatial-spectral reconstruction and k-t spirals for accelerated 2D spatial/1D spectral imaging of 13C dynamics. Magnetic Resonance in Medicine. 2014;71(4):1435–1445. doi: 10.1002/mrm.24796. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 22.Duyn JH, Yang Y, Frank JA, van der Veen JW. Simple Correction Method for k-Space Trajectory Deviations in MRI. Journal of Magnetic Resonance. 1998;132(1):150–153. doi: 10.1006/jmre.1998.1396. [DOI] [PubMed] [Google Scholar]
  • 23.Yang QX, Posse S, Le Bihan D, Smith MB. Double-sampled echo-planar imaging at 3 Tesla. Journal of Magnetic Resonance, Series B. 1996;113(2):145–150. doi: 10.1006/jmrb.1996.0167. [DOI] [PubMed] [Google Scholar]
  • 24.Hu X, Le TH. Artifact reduction in EPI with phase-encoded reference scan. Magnetic resonance in medicine. 1996;36(1):166–171. doi: 10.1002/mrm.1910360126. [DOI] [PubMed] [Google Scholar]
  • 25.Miller JJ, Lau AZ, Teh I, Schneider JE, Kinchesh P, Smart S, Ball V, Sibson NR, Tyler DJ. Robust and high resolution hyperpolarized metabolic imaging of the rat heart at 7 t with 3d spectral-spatial EPI. Magnetic Resonance in Medicine. 2015 doi: 10.1002/mrm.25730. DOI: 10.1002/mrm.25730. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 26.Skare S, Clayton DB, Newbould R, Moseley M, Bammer R. A fast and robust minimum entropy based non-interactive Nyquist ghost correction algorithm; Proceedings of the 14th Annual Meeting of ISMRM; Seattle, WA. 2006. Abstract 2349. [Google Scholar]
  • 27.Block KT, Frahm J. Spiral imaging: A critical appraisal. Journal of Magnetic Resonance Imaging. 2005;21(6):657–668. doi: 10.1002/jmri.20320. [DOI] [PubMed] [Google Scholar]
  • 28.Durst M, Koellisch U, Frank A, Rancan G, Gringeri CV, Karas V, Wiesinger F, Menzel MI, Schwaiger M, Haase A, Schulte RF. Comparison of acquisition schemes for hyperpolarised 13C imaging. NMR in Biomedicine 2015:n/a-n/a. doi: 10.1002/nbm.3301. [DOI] [PubMed] [Google Scholar]
  • 29.Tsao J. Ultrafast imaging: Principles, pitfalls, solutions, and applications. Journal of Magnetic Resonance Imaging. 2010;32(2):252–266. doi: 10.1002/jmri.22239. [DOI] [PubMed] [Google Scholar]
  • 30.Nishimura DG, Irarrazabal P, Meyer CH. A Velocity k-Space Analysis of Flow Effects in Echo-Planar and Spiral Imaging. Magnetic Resonance in Medicine. 1995;33(4):549–556. doi: 10.1002/mrm.1910330414. [DOI] [PubMed] [Google Scholar]
  • 31.Glover GH, Lee AT. Motion Artifacts in fMRI: Comparison of 2DFT with PR and Spiral Scan Methods. Magnetic Resonance in Medicine. 1995;33(5):624–635. doi: 10.1002/mrm.1910330507. [DOI] [PubMed] [Google Scholar]
  • 32.Arunachalam A, Whitt D, Fish K, Giaquinto R, Piel J, Watkins R, Hancu I. Accelerated spectroscopic imaging of hyperpolarized C-13 pyruvate using SENSE parallel imaging. NMR in Biomedicine. 2009;22(8):867–873. doi: 10.1002/nbm.1401. [DOI] [PubMed] [Google Scholar]

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.

Supplementary Materials

Supp Fig S1

RESOURCES