Improved measurements of the -decay response of liquid xenon with the LUX detector

DS Akerib, S Alsum, HM Araújo, X Bai, J Balajthy… - Physical Review D, 2019 - APS
DS Akerib, S Alsum, HM Araújo, X Bai, J Balajthy, A Baxter, P Beltrame, EP Bernard…
Physical Review D, 2019APS
We report results from an extensive set of measurements of the β-decay response in liquid
xenon. These measurements are derived from high-statistics calibration data from injected
sources of both H 3 and C 14 in the LUX detector. The mean light-to-charge ratio is reported
for 13 electric field values ranging from 43 to 491 V/cm, and for energies ranging from 1.5 to
145 keV.
We report results from an extensive set of measurements of the -decay response in liquid xenon. These measurements are derived from high-statistics calibration data from injected sources of both and in the LUX detector. The mean light-to-charge ratio is reported for 13 electric field values ranging from 43 to , and for energies ranging from 1.5 to 145 keV.