CLIN. CHEM. 33/4, 604-605 (1987)
Improved Radioimmunoassay of Melatonin in Serum
W. Sleghart E. Ronca, G. Drexier, and S. Karall
ton, MA) and incubated this with 150 tL of an antimelatonin antibody solution (WHB). After 1 h at 37 #{176}C,
the
samples were transferred to an ice bath and 400 p.L of icecold saturated ammonium sulfate solution was added. After
1 h at 4#{176}C,
samples were diluted with 3.5 mL of ice-cold,
70% saturated ammonium sulfate solution, then ifitered
through Whatman GF/B glass-fiber ifiters. We washed the
filters once with 3.5 mL of the 70% saturated ammonium
sulfate solution and then measured, by liquid scintillation
counting, the radioactivity remaining on the ifiters.
Additional Keyphrases:
samplepreparation
reversed-phase
circadianrhythms
The efficiency of the column extraction procedure was
assessed by adding 34 pg of [3H]melatomn to 0.6 mL of
serum and applying the mixture to Baker RP-18 columns.
After the columns had been centrifuged for 3 mm at 80 to
100 x g, >99% of the applied [3H]melatomn was retained by
the columns, only about 0.3% passing through them. In
other experiments not shown here we found that, even after
application of 5 mL of serum or 10 ng of melatonin, only
0.5% of the applied [3H]melatornn radioactivity passed
through the columns. More than 5 mL of serum tended to
clog the columns. We could prevent clogging by either using
adequate ifiter discs on the top of the columns, precipitating
the serum proteins before applying samples to the columns,
or digesting the serum proteins with trypsin.
Because the columns had to be washed to remove unbound proteins and lOWMr compounds in serum, we investigated various washing procedures that did not interfere
with binding of [3H]melatonin to the Baker RP-18 columns-e.g., up to five washes with 1 mL of distilled water or
one wash with 1 mL of pure n-hexane or 1 mL of aqueous
methanol (100 mJ.JL). In other experiments, columns were
washed with a 100 mLfL aqueous methanol that had been
adjusted with buffer to pH 3.0 or 9.0. Under all these
conditions >97% of the [3H]melatomn was still retained by
the columns. Because the concentrations of melatonin estimated with use of 100 mL/L methanol as a washing fluid
seemed to be slightly lower than those after the distilled
water washes, we used a single 1-mL wash with 100 mIJL
aqueous methanol for routine determination of melatomn in
serum.
Pure ethanol, methanol, or acetone completely eluted
melatonin from the columns. However, because of the
expense of pure ethanol and because a contaminant in
commercial acetone interferes with radioimmunoassay
of
melatonin, we routinely used 300 ,uL of methanol as eluent,
which rapidly and reproducibly eluted 97.4% (SD 0.8%) of
the applied [3H]melatonin. The used columns could be reused at least 10 times without further purification, if
pretreated with pure methanol and water as described
above in Materials and Methods.
To determine melatomn in the serum extract, we used a
method originally described by Wetterberg et al. (4) and
recommended by WFIB for use with their melatonin radioimmunoassay kit. We replaced the centrifugation with a
ifitration step involving Whatman GFIB glass-fiber filters.
This not only shortened the time necessary for the radioim-
.
chromatography,
Because of the extremely low concentrations of melatonin
in serum and the interference of serum components with the
radioiminunoassay for melatonin, most studies performed so
far have involved extraction of melatonin from serum before
analysis (1). However, the usual liquid-liquid
extraction is
not only incomplete but also time consuming. Therefore, we
investigated the usefulness of a rapid column-extraction
method, such as has been developed for various other
compounds (2, 3), for the extraction of melatonin. In addition, we modified a commercially available radioimmunoassay for melatonin to improve reproducibility and
simplicity.
Materials and Methods
To extract the melatonin, we used Baker-lO SPET’ RP-18
columns (J. T. Baker Chemical Co., Phillipsburg, NJ 08865)
pretreated by washing twice with 1 mL of pure methanol,
and then twice with 1 mL of distilled water. These washing
steps are absolutely essential for the correct function of the
columns. After applying 0.1 to 5 mL of serum, we washed
the columns with 1 mL of aqueous methanol (methanoll
water, 10/90 by vol), then eluted melatonin from the
columns by applying 300 L of pure methanol. The spontaneous flow of solvents or serum through the columns was
rather low, but when we tried to speed the flow by applying
suction, as recommended in the respective extraction protocols (2, 3), the proportion of melatonin extracted varied.
Because it was essential to keep the columns wet throughout the washing and extraction procedure, we routinely
centrifuged the columns at 80 to 100 x g for 3 miii to
improve the flow rate of solvents or serum through them.
After evaporating the methanol from the samples we
determined the melatonin by using a modification of a
commercially
available
radioimmunoassay
(WHB,
Bromma, Sweden). We redissolved the melatonin residue in
250 iL of sodium phosphate buffer (50 mmol/L, pH 7.5)
containing, per liter, 154 mmol of NaC1, 15 mmol of sodium
azide, 1 g of gelatin, 1 g of gamma-globulin, and 0.25 nmol of
[3H]melatonin (76.3 kCilmol; New England Nuclear, BosPsychiatrische Universitataklinik, Allgemeines Krankenhaus
der Stadt Wien, Wahringer GUrtel 18-20, 1090 Vienna, Austria.
Received September 29, 1986; accepted January 27, 1987.
604 CLINICALCHEMISTRY, Vol. 33, No. 4, 1987
Results
Downloaded from https://academic.oup.com/clinchem/article/33/4/604/5653239 by guest on 29 November 2021
Melatonin was extracted from serum by using Baker reversed-phase C-i 8 columns. More than 99% of the applied
melatonin was retained by the columns, and more than 97%
was eluted from the columns in 300 pt of methanol. We then
determined melatonin in the serum extract by a modification
of a standard radioimmunoassay, using filtration instead of
centrifugation to collect the [3H]melatonin-antibody complex
precipitated by saturated ammonium sulfate. These modifications allow more rapid, accurate, and reproducible determination of melatonin than do previously published procedures.
Table 1. Accuracy and Precision of the Modified R1Afor
Determining Concentrations of Melatonin In Seruma
Concn In
serum,
nWL
22.4
34.9
62.4
102.4
Within run (n
=
4)
Mean,
22.0
33.5
64.5
108.0
First concentration
Mean,
CV, %
ng
Betweenrun(n = 11)
8.8
8.5
6.8
5.0
is for unsupplemented
nL
22.4
34.0
66.1
106.0
CV, %
11.6
11.8
7.6
serum, others had added
melatonin.
Discussion
This rapid column-extraction method for extracting melatonin from serum is very reproducible and yields a high
proportion of applied melatonin in the column eluate. In
easily evaporated than the relatively large volumes involved in liquid-liquid
extraction
of this compound. By
using the described method, one can extract 40 samples
within 90 miii and assay them within 3 to 4 h, with better
reproducibility and accuracy than for other melatomn determinations. Thus, it is easily possible to assay melatonun in
40 samples every day.
References
1. Arendt J. Assay of melatonin and its metabolites: results in
normal and unusual environments. In: Wurtman R, Waldhauser F,
eds. Melatomn in humans (Proc. First hit. Conf. on Melatonin in
Humans). Vienna, Austria: Center for Brain Sciencesand Metabolism Charitable Trust, 1985:11-32.
2. Good TJ, Andrews JS. The use of bonded-phase extraction
columns for rapid sample preparation of benzodiazepines and
metabolites from serum for HPLC-analysis. J Chromatogr Sci
1981;19:562-6.
3. Application Note for Bond Elut#{174}
extraction methods. AnalytichemInternational, Harbor City, CA 90710.
4. Wetterberg L, Eriksson 0, Friberg Y, Vangbo B. A simplified
radioimmunoassay for melatonin
and its applicationto biological
fluids. Preliminary observations on the half life of plasma melatonm. Clin Chim Acta 1978;86:169-77.
CLINICALCHEMISTRY, Vol. 33, No. 4, 1987 605
Downloaded from https://academic.oup.com/clinchem/article/33/4/604/5653239 by guest on 29 November 2021
munoassay but also decreased blank values, thus improving
the accuracy and reproducibility
of the assay. Using this
extraction and radioimmunoassay procedure for melatornn,
we recovered between 96 and 104% of applied melatonin in
various experiments, independently of the volume of serum
or buffer applied to the columns. The intra-assay CV was
between 5 and 9%, the interassay CV between 5 and 12%
(Table 1).
Using the described method, we investigated circadian
rhythms of melatomn concentrations in serum from 15
patients of the Psychiatric University Clinic in Vienna
(experiments to be published elsewhere). The lowest concentrations (by day) were 5 to 21 pg/mL (mean ± SD: 13.5 ± 4
pg/mL), and the greatest values (during the night) were 31
to 180 pg/mL (86 ± 45 pglmL).
addition, this extraction procedure is versatile. It can easily
be modified by using various solvents or buffer solutions as
washing media for removing from the columns compounds
that might interfere with subsequent assay of melatonin.
This column-extraction method thus might be beneficial for
the quantification
of melatonin, not only by radioimmunoassay but also by gas chromatography-mass
spectrometry or by liquid-chromatographic
methods.
Finally, the relatively small volume of organic solvent
needed to elute the melatonin from the columns can be more