NIM B
Beam Interactions
with Materials & Atoms
Nuclear Instruments and Methods in Physics Research B 243 (2006) 392–396
www.elsevier.com/locate/nimb
Simple and accurate spectra normalization in ion beam analysis
using a transmission mesh-based charge integration
M. El Bouanani a, P. Pelicon b,*, A. Razpet b, I. Čadež b, M. Budnar b,
J. Simčič b, S. Markelj b
a
b
Laboratory for Electronic Materials and Devices, Department of Materials Science and Engineering, University of North Texas,
Denton, TX 76203-5310, USA
Microanalytical Center, Department of Low and Medium Energy Physics, Jožef Stefan Institute, P.O. Box 3000, SI-1001 Ljubljana, Slovenia
Received 27 December 2004; received in revised form 25 July 2005
Available online 2 November 2005
Abstract
Accurate and reproducible determination of the number of impact ions is essential for quantitative IBA measurements. Herewith we
present an in-beam charge-collection device, consisting of a tungsten mesh enclosed by two negatively biased annular electrodes and a
shaping slit. The charge-collection efficiency was measured as a function of aperture bias and the reproducibility of charge collection at
different bias voltages studied. Scanning transmission ion microscopy (STIM) was used to check the effect of beam scattering at the mesh
on its energy distribution. The effect of the device on the primary beam energy distribution was calculated for the beams typically used in
Rutherford backscattering spectroscopy (RBS) and elastic recoil detection analysis (ERDA). Excellent characteristics of the device were
demonstrated.
2005 Elsevier B.V. All rights reserved.
PACS: 82.80.Yc; 68.49.Sf
Keywords: Beam charge normalization; Rutherford backscattering spectroscopy (RBS); Elastic recoil detection analysis (ERDA); Ion implantation
1. Introduction
Highly accurate measurement of the number of impact
ions during the acquisition of spectra is a key issue for
the precision of ion beam analysis (IBA). IBA beam normalization methods could be classified according to their
principle of operation as follows:
A: Charge integration
• Total beam charge integration from an isolated chamber
enclosing the sample. The chamber acts a Faraday cup.
*
Corresponding author. Tel.: +386 1 5885 294; fax: +386 1 5885 377.
E-mail address: primoz.pelicon@ijs.si (P. Pelicon).
URL: http://www.rcp.ijs.si/~mic/ (P. Pelicon).
0168-583X/$ - see front matter 2005 Elsevier B.V. All rights reserved.
doi:10.1016/j.nimb.2005.09.002
• Total charge integration from a positively biased sample
for suppression of secondary electrons [1].
• In-beam charge integration devices that intercept a fraction of the primary ion beam. Integrated charge is proportional to the accumulated charge on the sample [2].
B: Secondary particle and photon emission
• From the sample.
• From in-beam chopping devices [3].
• From in-beam media, i.e. X-ray yield from argon or an
exit foil used in in-air proton-induced X-ray emission
(PIXE) analysis.
The methods of the first group (A) use charge integrators for determination of the number of ions impinging
on sample during the measurement. The process of
M. El Bouanani et al. / Nucl. Instr. and Meth. in Phys. Res. B 243 (2006) 392–396
secondary electron emission, which increases the positive
current readout at the irradiated sample, is strongly dependent on the beam-interacting surface and the intensity
of local electric and magnetic fields. The stability of
charge integration during a particular measurement may
also be affected due to the beam-induced changes at the
irradiated sample surface. The known effect of beamassisted hydrocarbon deposition at the beam-interacting
surface strongly depends on the vacuum conditions. The
reproducibility of the method therefore depends on the efficiency of secondary electron suppression. The normalization methods based on charge integration (group A) at
beam currents below 1 nA, as is the case for the ion currents in nuclear microprobes, may yield significant error
due to contribution of stray currents.
The methods of the group B are in general less restricted
regarding the beam intensity. The accumulation rate of
normalizing events should be matched to a particular
application in order to avoid excessive statistical error or
significant dead time.
The type of device discussed in this work belongs to the
class of in-beam charge-collection devices (group A) and
has been, according to our knowledge, suggested and
implemented for standard IBA characterization for the first
time at the University of North Texas (UNT) by El Bouanani. The original device in use at UNT is made of a tungsten mesh that is completely enclosed in negatively biased
cylinder electrodes with an entry and exit aperture slits
attached. A fine mesh on the annular holder, surrounded
by the negatively biased cylinders, is used to collect fraction
of charge that is proportional to the total charge of the
transmitted beam. The entire transmitted beam through
the mesh charge-collection device then passes to the sample
without further obstructions. Due to high wire density, the
used mesh takes into account the beam intensity variations
over the entire beam cross-section.
2. The mesh based charge-integration device
The schematic drawing of the mesh charge-integration
device constructed and in use at Jožef Stefan Institute
(JSI) is shown in Fig. 1. The annular electrodes are separated by insulating ruby balls. The walls of the electrodes
are shaped to shield the insulating ruby balls from the
beam trajectory to avoid their charging. The first electrode
in the stack is grounded and contains the mounting for the
replaceable tantalum beam-shaping slit. The mesh is
mounted between two annular elements. It is encompassed
by two negatively biased electrodes to suppress the emission of secondary electrons emitted from the irradiated
area of the mesh. Two electric connections are led through
coaxial vacuum feedthroughs: the mesh contacts the charge
integrator and the bias contacts the negative pole of the
DC voltage power supply.
The mesh in use consists of woven tungsten wires each
with a diameter of 38 lm (0.0015 in.) and a density of
3.2 lines/mm (80 lines/in.) with 77.4% transmission.
The mesh charge-integration device is mounted at the
entrance of the RBS/ERDA scattering chamber at
the 10 beamline of the 2 MV JSI tandem accelerator.
The beam profile along the beamline is defined by an
electrostatic quadrupole lens which is positioned at the
accelerator exit. By means of this lens, the beam is focused
on the shaping slit of the mesh charge-integration device.
The shape and the size of the slit is chosen to obtain a desirable beam spot on the sample. In the case of simultaneous
30 mm
8 mm
BEAM
Replaceable
shaping slit
Insulating ruby
balls
393
Charge int.
Neg. bias
Fig. 1. Construction of the in-beam mesh charge collector. The tungsten mesh is inserted between the two central electrodes.
394
M. El Bouanani et al. / Nucl. Instr. and Meth. in Phys. Res. B 243 (2006) 392–396
3. Effect of bias on the charge-collection stability and
efficiency
The effect of the bias voltage applied to the electrodes
enclosing the mesh on the charge-collection efficiency
(Eq. (1)) was measured using the RBS signal from a sample
consisting of a thin film of TaN on Si [4] with a known Ta
areal density. A 4.2 MeV 7Li2+ beam was used [5] and the
integrated number of counts in the tantalum peak per constant integrated charge on the mesh was measured. Using
the same sample and detector setup, we performed a series
of measurements under different beam intensities in
different runs over a period of one month. Three different
bias voltages were evaluated. With a negative bias voltage
higher than 500 V the charge integration was extremely
reproducible, which is not the case for the low negative bias
of 20 V (Fig. 2). Similar tests with the device built at UNT
and in use at the UNT RBS setup demonstrated the reproducibility better than 1% with the He+ beam using the energies between 1 and 2 MeV.
Further RBS measurements at JSI on the TaN/Si
sample revealed a strong dependence of the Ta RBS yield
per unit of mesh-integrated charge on the bias voltage
(Fig. 3). To describe the observed results, the meshintegrated charge Qint can be presented as a sum of the
following contributions:
Qint ¼ AN beam Ze0 þ MAN beam e0 SN beam e0 .
ð1Þ
No. of counts/0.1 microcoulomb
The quantity A is the relative area consisting of the mesh
wires that blocks the beam and equals 1 T, where T
stands for the mesh transmission. For this particular study
A equals 0.226. Ze0 is the charge of the ions in the beam
18000
16000
14000
20 V
4
3
Charge collection efficiency
measurement of RBS and ERDA spectra, the sample is
tilted by 75 angular degrees and the shaping slit used is
1 mm wide and 4 mm high. This yields a square-like beam
spot on the sample with the approximate size of 4 · 4 mm2.
2
1
1
10
100
1000
Negative bias [V]
Fig. 3. Measured charge integration efficiency D (Eq. (2)) as a function of
the negative bias at the suppression electrodes. The error bars represent
the statistical uncertainty.
used for the analyses. The first term in the sum (Eq. (1))
is the charge integrated from the stopped fraction of the
primary beam consisting of Nbeam ions. The second term
represents the effective positive charge collected from the
mesh due to the escape of secondary electrons, where M describes the number of emitted secondary electrons per ion
hitting the mesh. The third, negative term accounts for
the contributions of positive ions emitted from the mesh
and electrons hitting the mesh. Both contributions are
assumed to be proportional to the number of ions in the
primary ion beam Nbeam with a proportionality constant
Se0. Fig. 2 shows the measured charge-collection efficiency
D as a function of bias voltage and is defined by
D ¼ Qint =ðAN beam Ze0 Þ;
ð2Þ
as a function of bias voltage. The sharp decrease of collection efficiency in the negative bias region from 0 to 10 V is a
result of strong suppression of slow secondary electrons.
Suppression of the processes described by the third term
in Eq. (1) is achieved with a much higher negative bias.
At negative bias voltages over 600 V the charge collection
efficiency does not depend on the bias voltage. In this region, the contributions from both the second and third
term in Eq. (1) are completely suppressed and the measured
charge corresponds to the charge of the ions stopped in the
mesh.
300 V
800 V
4. Application of the mesh charge-integration device for RBS
and ERDA spectra normalization
12000
10000
Consecutive runs
Fig. 2. The stability of the device checked at bias voltages of 20, 300
and 800 V. The number of counts in the tantalum peak has been
measured in the RBS spectrum of a thin film of TaN on Si. Error bars
show one standard deviation of the number of accumulated counts.
The normalization parameter P of the RBS/ERDA
measurement, given as the number of ions hitting the sample during measurement Ni multiplied by the detector solid
angle Xdet, is proportional to the integrated charge Qint on
the mesh electrode:
P ¼ N i Xdet ¼ K Qint .
ð3Þ
395
M. El Bouanani et al. / Nucl. Instr. and Meth. in Phys. Res. B 243 (2006) 392–396
The proportionality constant K is determined by RBS
measurements on standard reference samples. Thin film
reference standards for RBS with a known area density
of a deposited heavy element on a substrate consisting of
a light element are the best choice. Thin film reference
standards with a certified precision of better than 2% are
available [6,7]. An alternative to these standards are thick
pure amorphous targets with very well known stopping
powers for the ions in use (e.g. 4He ions in Au).
The geometrically determined value of the solid angle
Xdet is used at the JSI setup only to obtain an approximate
value for the number of impact ions Ni during measurement from the normalization parameter P. The resulting
Ni is used for evaluation of beam damage effects during
the measurement. Use of the number of impact ions Ni
for normalization of the RBS/ERDA analysis would introduce an additional experimental uncertainty to the results.
5. Effect of the mesh on the primary ion beam
The model presented in Section 3, which describes the
charge-collection efficiency of the device, neglects processes
at the mesh edges, where a part of the beam is subject to
only partial stopping. This is expected to affect both the
energy and the angular distribution of the primary beam.
To estimate the fraction of the beam which is subject to
partial stopping at the edges of the mesh wire, the finite
range of the ions in the mesh material should be considered. If the radius of the wire is r and the range of the
ion in the mesh material is R, then the radial distance p
from the edge at which the ion could penetrate the wire is
qffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffi
2
p ¼ r 1 ð1 ðR=ð2rÞÞ Þ .
ð4Þ
For the square mesh with the wire distance of D, the fraction of the beam at the sample, which undergoes the deceleration process in the mesh wires and has non-zero energy,
is given by
F ffi
4pðD 2r þ 2pÞ
2
ðD 2r þ 2pÞ
.
F ffi
4
ðp=rÞ.
ðD=rÞ 2
ð6Þ
Table 1 lists the calculations of the contaminated fractions
of the incident beam based on this geometrical model (Eqs.
4 and 6) for some ion beam types and energies typically
used in IBA analysis. The set of beams evaluated includes
2 MeV protons, typically used in proton-induced X-ray
emission (PIXE), 2 MeV He ions frequently used for
RBS and hydrogen-ERDA and beams used for heavy-ion
ERDA. As expected, the contaminated fraction of the incident beam by the mesh increases with the increase in the
range of ions in the mesh material. The case of 2 MeV protons exhibits the most significant fraction of contamination
among the ion beams included in Table 1. For the other ion
beams than protons that are typically used in IBA analyses,
the contaminated fraction is in the order of one ion per
thousand or lower.
Direct measurement of the energy distortion of the incident ion beam through the tungsten mesh was performed
using scanning transmission ion microscopy (STIM). A
2 MeV proton beam with a diameter of 1 lm and an intensity of ca. 100 ions/s was used for the STIM measurement.
An area of 1 · 1 mm2 was uniformly scanned and the ion
detector was positioned directly in the beam. Measured
spectra of the beam with and without the mesh are shown
in Fig. 4. The peaks in both spectra fit well a Gaussian distribution with a width r of 16.7 keV. To evaluate the energy
distortion of the beam that is due to the mesh, examination
of the collected STIM spectra without the mesh was first
undertaken. The integrated yield I is defined as follows:
Z E0 5r
I¼
ðdn=dEÞ dE;
ð7Þ
0
where E0 is the energy of the primary beam and the (dn/dE)
refers to the measured STIM spectrum. The beam
empty space
ð5Þ
mesh in the beam
1000
Table 1
Range R of ions in tungsten [8], penetration range given as the ratio p/r
(Eq. (4)) from the edge of the tungsten wire in the mesh used (radius
19 lm) and the fraction of the beam F with altered energy due to
positioning of the mesh in the beam (Eq. (6))
Beam
R (lm)
p/r (Eq. (4))
F (Eq. (6))
1
14.3
4.25
3.5
0.98
3.1
3.8
4.6
7.4e2
2.5e3
4.25e3
3.3e4
3.3e3
5.0e3
7.3e3
2.0e2
6.9e4
1.1e3
9.1e5
9.1e4
1.4e3
2.0e3
H, 2 MeV
He, 2 MeV
7
Li, 4.2 MeV
12
C, 2 MeV
19
F, 14.2 MeV
35
Cl, 35 MeV
127
I, 70 MeV
4
Linear trajectories of ions in the mesh material are assumed.
Yield
Since Eq. (5) stands for all the relevant cases where p/r 1
(Table 1), it could be rewritten as
100
10
1
100
200
300
Channel
Fig. 4. Effect of the mesh on the energy distribution of the primary beam:
a 2 MeV proton microbeam was scanned over the mesh area of 1 · 1 mm.
The detector for scanning transmission ion microscopy (STIM) was
positioned at 0. The result of the blank scan is included for comparison.
396
M. El Bouanani et al. / Nucl. Instr. and Meth. in Phys. Res. B 243 (2006) 392–396
contamination C is defined as measure of the ratio of I and
the total number of detected ions Itot:
C ¼ I=I tot .
ð8Þ
The spectrum of the blank scan revealed contamination of
2%, which is result of the extremely small object slit
opening during the 0 STIM measurement. For the STIM
spectrum of the mesh, a weak low energy tail is present in
the spectrum and the fraction of contamination increases
to 4%. This rise of contamination of 2% is in reasonable
agreement of the prediction of the geometrical calculation
(Table 1).
The 2 MeV proton beam, studied here by STIM, is
expected to exhibit the highest fraction of contamination
among the beam types discussed in Table 1. The use of
He ions or heavy ions results in a significantly lower beam
contamination, which enables high sensitive ion beam
analysis [5,9].
analysis (ERDA) and nuclear reaction analysis (NRA).
The presented results encourage the use of the device also
at the field of ion implantation. The sweeping of the beam,
which is usually applied in the implantors to achieve the
fluence uniformity over the large implanted areas, should
be applied in this case after the beam final shaping and
charge collection with the presented device in order to
avoid shade effects.
Acknowledgement
This work at JSI was supported by the grant J1-32790106-01 from the Slovenian Ministry of Education,
Science and Sport, by the IAEA Co-ordinated Research
Project ‘‘Analysis of light elements in thin films, including
depth profiling’’ and the bilateral project SLO-USA-2002/
19.
6. Conclusion
References
The proposed mesh-based charge integration device
in combination with known thin film standards is a simple
and very effective solution for the accurate spectra
normalization in ion beam analysis. It features reproducibility of better than 1%, exceptional stability and outstanding reliability when compared to the instrumentally
more complex transmission Faraday cup [2]. The effects
of the mesh on the primary incident beam were carefully
evaluated and it was experimentally demonstrated that
the distortion is negligible for most of the standard beams
used in conventional ion beam analysis except for protons,
where a thicker tungsten mesh would perform better than
the one studied here. Its ease of construction and simple
use provide great potential for applications in Rutherford
backscattering spectroscopy (RBS), elastic recoil detection
[1] F. Herrmann, D. Grambole, Nucl. Instr. and Meth. B 104 (1995)
26.
[2] F. Pászti, A. Manuaba, C. Hajdu, A.A. Melo, M.F. Da Silva, Nucl.
Instr. and Meth. B 47 (1990) 187.
[3] L. Bartha, I. Uzonyi, Nucl. Instr. and Meth. B 161–163 (2000) 339.
[4] W.A. Lanford, P. Pelicon, B. Zorko, M. Budnar, Nucl. Instr. and
Meth. B 190 (2002) 410.
[5] P. Pelicon, A. Razpet, S. Markelj, I. Čadež, M. Budnar, Nucl. Instr.
and Meth. B 227 (2005) 591.
[6] K.H. Ecker, A. Berger, R. Grotzschel, L. Persson, U. Watjen, Nucl.
Instr. and Meth. B 175 (2001) 797.
[7] G. Boudreault, C. Jeynes, E. Wendler, A. Nejim, R.P. Webb, U.
Watjen, Surf. Interf. Anal. 33 (2002) 478.
[8] J.F. Ziegler, SRIM program. Available from: <http://www.srim.org/>.
[9] P. Pelicon, G.V. Ravi Prasad, M. El Bouanani, B.N. Guo, D. Birt,
J.L. Duggan, F.D. McDaniel, in: Proceedings of 17th International
Conference on the Application of Accelerators in Research and
Industry, Denton, 2002, AIP Conf. Proc. 680 (1) (2003) 482.