REVIEW OF SCIENTIFIC INSTRUMENTS
VOLUME 70, NUMBER 7
JULY 1999
Deconvolution and measurement of bulk and surface optical absorptions
in Ti:Al2O3 crystals using photopyroelectric interferometry
Chinhua Wang and Andreas Mandelisa)
Photothermal and Optoelectronic Diagnostics Laboratories, Department of Mechanical and Industrial
Engineering, University of Toronto, Toronto M5S 3G8, Canada
~Received 19 January 1999; accepted for publication 25 March 1999!
The extension of our earlier single-layer ~monolithic! photopyroelectric ~PPE! interferometric
theory to include surface and bulk optical absorptions has allowed the measurement of both bulk
absorption coefficient and surface absorptance in one single experiment. Based on purely
thermal-wave interferometry, the thermal-wave cavity lengths of a PPE interferometer were scanned
using pairs of Ti: sapphire crystals with appropriate combinations of figure of merit, surface polish,
and thickness. In the conventional single-ended ~noninterferometric! PPE technique, the surface
reflectivity, surface absorptance, and bulk absorption coefficient are always coupled together.
However, PPE destructive interferometry provides a method for extracting highly precise values of
one of these optical parameters, without the requirement of equally precise knowledge of the values
of the others. © 1999 American Institute of Physics. @S0034-6748~99!02007-9#
I. INTRODUCTION
sample under investigation. With regard to transparent materials, this large baseline signal, which appears in the lock-in
IP channel, is much larger than that of the Q channel, by
more than two orders of magnitude.3 As a result, one must
choose a very low instrumental sensitivity to prevent the
lock-in amplifier from overloading due to this large baseline
signal. Usually, this operation limits the dynamic range of
the PPE measurement, which is far too low to detect small
changes in both IP and Q channels where high quality laser
crystals are concerned, with minute differences in optical
properties. Moreover, in comparing different crystals, it is
difficult to ensure identical alignment procedures for the two
individual measurements. Regarding accuracy, the PPE measurement of b~l! in the single-ended purely optical transmission mode is actually a single-point measurement at a specific wavelength. The one-point measurement thus lacks the
accuracy afforded by linear correlation and averaging process, and the measurement depends directly on other optical
parameters, such as surface reflectivity of the sample. Very
recently, a PPE interferometric technique was introduced and
applied to the coherent suppression of the large baseline signal for the characterization of transparent samples, as well as
for the study of the thermophysical properties of intracavity
gases.17–19 The figure of merit for the large baseline suppression, defined as the single-ended-to-interferometric signal
amplitude ratio, was found to be over 33103 in practice,
with a 5 mW laser source.18 The measurement precision,
signal dynamic range and signal-to-noise ratio ~SNR! were
also much improved over the single-ended configuration
with the thermal-wave interferometric technique. One of the
earlier experimental results shows the difference in total
~bulk and surface! optical absorption coefficients between a
pair of nominally identical BK7 window glasses using our
monolithic PPE interferometric theory.18 It is believed, however, that this difference in the total absorption coefficient is
mainly due to the difference between the surface absorptan-
Photothermal ~PT! and photoacoustic ~PA! spectroscopic
techniques have been used successfully in measurements of
optical absorption coefficients and nonradiative quantum efficiencies for a variety of optical materials, including laser
crystals.1–13 Among the various embodiments of PA and PT
techniques, photopyroelectric detection ~PPE!, as pointed out
elsewhere,1–4,10–12,14 has a certain number of advantages. A
major advantage of PPE detection over other conventional
PT techniques is the fact that one can measure directly and
self-consistently both the optical absorption coefficient, b~l!,
and the nonradiative quantum efficiency, h~l!.2,3,12,15,16
Lock-in quadrature photopyroelectric spectroscopy ~QPPES! was used in a novel noncontact experimental scheme
to obtain high-resolution spectra of the nonradiative quantum
efficiency of Ti:sapphire laser crystals with widely different
figures of merit ~FOM!.2 Using the same setup, opticalabsorption-coefficient spectra were obtained from the lock-in
in-phase ~IP! channel in a separate measurement, in the socalled ‘‘purely optical transmission ~OT! mode.’’ The singlelayer ~monolithic!, single-ended theoretical model,2 however, allowed only the determination of total absorptance
~bulk and surface! values. Subsequently, Vanniasinkam
et al.3 separated the surface absorptance from the bulk absorption coefficient in a modified PPE theory. Those authors
employed two samples of identical bulk quality and surface
treatment, but different thicknesses and performed two independent PPE experimental measurements in the conventional
single-ended configuration. The contributions from the surface and the bulk were thus separated from each other. It was
shown,3,11,12 however, that the sensitivity of conventional
single-ended PPE measurements suffers from a large baseline signal, generated by the direct transmission to the detector surface of the incident radiation through the transparent
a!
Electronic mail: mandelis@mie.utoronto.ca
0034-6748/99/70(7)/3115/10/$15.00
3115
© 1999 American Institute of Physics
Downloaded 18 Jul 2008 to 128.100.49.17. Redistribution subject to AIP license or copyright; see http://rsi.aip.org/rsi/copyright.jsp
3116
Rev. Sci. Instrum., Vol. 70, No. 7, July 1999
C. Wang and A. Mandelis
dimensional heat diffusion equations subject to appropriate
boundary conditions of thermal-wave field continuity and
flux conservation across each interface ~g1-s, s-g2, g2-p,
p-g3, g3-r, and r-g4! of Fig. 1. Assuming the whole system is exposed to the same gaseous atmosphere ~usually,
air!, the appropriate thermal-wave equations have the form:
d 2T i~ x !
2 s 2i T i ~ x ! 50,
dx 2
FIG. 1. Schematic of the photopyroelectric interferometric setup for the
theoretical analysis. Bulk optical absorption coefficients of sample and reference, respectively: b s (l), b r (l); nonradiative energy conversion efficiency of sample and reference, respectively: h sb (l), h rb (l); surface absorptance of sample and reference, respectively: A s ,A r ; thickness of sample and
reference, respectively: l,m; thickness of PVDF detector: d; sample-PVDF
and PVDF-reference cavity lengths, respectively: L,L 1 .
ces of the two samples as a result of the antireflection coating
optical properties rather than due to the difference in bulk
optical properties. In addition, investigations which have
studied the impact of the quality of laser crystal rods on
output efficiency10,20,21 have demonstrated an intimate correlation between the slope efficiency and lasing threshold and
the quality of the laser crystal, including bulk crystal preparation and surface polish. Therefore, the requirement for improved control and evaluation of bulk optical properties of
laser crystals as a function of preparation, as well as the
effects of surface polish, motivates the deconvolution of surface absorptance from that of the bulk using our novel PPE
interferometric technique. In this article, the PPE interferometric theory18 is extended to separately include the surface
and bulk absorption in an optical material. The extended
theory is then applied to high-precision measurements performed with various combinations of Ti sapphire laser crystals with different figures of merit, polishes, and thicknesses.
for regions g1, g2, g3, g4, and p,
~1a!
i51,2,3,4,p
d 2T s~ x !
I ts ~ x !
2
b
,
2 2 s s T s ~ x ! 52 h s b s
dx
2k s
~1b!
d 2T r~ x !
I tr ~ x !
2
b
,
2 2 s r T r ~ x ! 52 h r b r
dx
2k r
l1L1d1L 1 <x<l1L1d1L 1 1m.
~1c!
In Eqs. ~1!, s i 5(11 j) Av /2a i is the complex thermal diffusion coefficient in spatial region i (i5g1,g2,g3,g4,s, p,r)
with thermal diffusivity a i ; k s ,k r is the thermal conductivity
of the sample and the reference, respectively; I ts (x),I tr (x)
are the total optical fluence contributions to depth x in the
sample and in the reference, respectively.2 They have been
derived upon considering the multiple reflections of the incident and the reflected light by the metal electrode~coating! of
the detector back into the sample and the reference, and are
given as follows:
I ts ~ x ! 5I 1
~ 12R s ! e 2A s
12R 2s e 22 ~ b s l1A s !
~ N 1 e 2 b s x 1N 2 e 2 b s ~ 2l2x ! ! ,
~2a!
I tr ~ x ! 5I 2 e j w
II. THEORY
In the interferometric configuration of Fig. 1, two laser
beams of intensities I 1 and I 2 , respectively, are split off from
a laser source and are modulated at the same angular frequency ~v!. They have a fixed, adjustable phase shift ~Dw!,
and are incident onto the front and rear surfaces of a polyvinylidene fluoride ~PVDF! thin film detector, passing through
optically transparent sample and reference media, which,
along with the PVDF sensor in the middle form the thermalwave cavities g2 and g3 as shown in Fig. 1. The PVDF
detector used in this work has a thickness of 52 mm and is
coated with Ni–Al alloy on both sides acting as electrodes.
The thickness of each electrode is about 80 nm, thus, it is
nearly opaque at 632.8 nm of incident laser light. In our
calculations, the optical absorption coefficient of the electrodes was considered infinite. The incident beams are assumed to illuminate the PVDF sensor uniformly with spotsizes much larger than the thermal diffusion length in PVDF,
so that the one dimensionality of the heat transfer model is
assured. The photopyroelectric signal from the PVDF detector is proportional to the average ac temperature of the
PVDF film detector.1 It is governed by coupled one-
0<x<l,
~ 12R r ! e 2A r
12R 2r e 22 ~ b r m1A r !
3 ~ N 1r e 2 b r ~ l1L1d1L 1 1m2x !
1N 2r e 2 b r @ 2m2 ~ l1L1d1L 1 1m2x !# ! .
~2b!
Here, R s ,R r ,R p are the surface reflectances of the sample, of
the reference, and of the PVDF detector, respectively. N 1 ,N 2
and N 1r ,N 2r are combinations of constants including surface
absorptance, bulk absorption coefficient, sample and reference thickness. The detailed expressions are given in the
Appendix. The solutions of the foregoing thermal-wave
equations contain coupled constants via the boundary conditions of temperature continuity ~absence of interfacial thermal resistance! and heat flux discontinuities at four surfaces
of the sample and the reference ~presence of infinitesimal
thin absorbing layers acting like interfacial sources!. To formulate the heat flux conservation relations, the thermal-wave
Eqs. ~1b! and ~1c! are integrated over a surface layer of
thickness e →0 at each of four surfaces of the sample and the
reference,3 respectively:
kg
dT 1 ~ x !
dT s ~ x !
2k s
5 h ~s0 ! A s I t ~ 0 ! ,
dx
dx
x50,
~3a!
Downloaded 18 Jul 2008 to 128.100.49.17. Redistribution subject to AIP license or copyright; see http://rsi.aip.org/rsi/copyright.jsp
Rev. Sci. Instrum., Vol. 70, No. 7, July 1999
ks
kg
dT g ~ x !
dT s ~ x !
2k g
5 h ~s0 ! A s I t ~ l ! ,
dx
dx
dT 3 ~ x !
dT r ~ x !
2k r
5 h ~r0 ! A r I tr ~ m ! ,
dx
dx
Bulk and surface optical absorptions
x5l,
~3b!
x5l1L1d1L 1 ,
3117
where e is the effective thickness of the surface layer. Equations ~3a!–~3d! replace the thermal flux continuity boundary
conditions at material surfaces used in Ref. 18. The thermal
flux discontinuities at the surfaces of the pyroelectric detector are
~3c!
kr
dT 3 ~ x !
dT g ~ x !
2k g
5 h ~r0 ! A r I tr ~ 0 ! , x5l1L1d
dx
dt
1L 1 1m,
~3d!
where I t (0), I t (l), I tr (0), and I tr (m) are constants related to
b s , b r ,A s ,A r . They are given in the Appendix. At each surface of the sample and the reference, the absorptance and the
nonradiative energy conversion efficiency is defined as
A i [lime →0 ~ b i e ! ,
i5s,r
~4!
h ~i 0 ! [lime →0 h ~i e ! ,
i5s,r,
~5!
and
V~ v !5
kg
dT 2 ~ x !
dT p ~ x !
2k p
5 ~ 12R p ! I 1 ~ l ! ,
dx
dx
x5l1L, ~6a!
kp
dT p ~ x !
dT 3 ~ x !
2k g
5 ~ 12R p ! I 2 ~ m ! ,
dx
dx
x5l1L1d,
~6b!
where I 1 (l) and I 2 (m) represent the incident beam intensities at the front and rear surfaces of the PVDF thin film,
respectively. They are given in detail in the Appendix. Following the algebraic procedure of Ref. 18, we finally obtain
a modified expression for the photopyroelectric interferometric signal associated with the geometry of Fig. 1:
S~ v !
H 1 G 1 ~ 11W 21e 22 s g L ! 1H 2 G 2 ~ 11V 34e 22 s g L 1 ! 12b gp ~ H 1 G 3 e 2 s g L 1H 2 G 4 e 2 s g L 1 !
.
3 s pd
s p ~ 11b gp ! e ~ 11 g gp V 34e 22 s g L 1 !~ 11 g gp W 21e 22 s g L ! 2e 2 s p d ~ g gp 1W 21e 22 s g L !~ g gp 1V 34e 22 s g L 1 !
S( v ) is the instrumental transfer function. It can usually be
normalized out experimentally. In addition,
b i j 5k i Aa j /k j Aa i ,
~8a!
g i j 5 ~ 12b i j ! / ~ 11b i j ! ,
~8b!
W 2152 g gs
V 3452 g gr
e s s l 2e 2 s s l
e
ssl
2 g 2gs e 2 s s l
~8c!
,
e s r m 2e 2 s r m
e s r m 2 g 2gr e 2 s r m
,
~8d!
G 15
~ 12R p ! I 1 ~ 12R s ! 2 e 2 ~ b s l12A s !
3
,
k ps p
12R 2s e 22 ~ b s l1A s !
~8e!
G 25
~ 12R p ! I 2 e jD w ~ 12R r ! 2 e 2 ~ b r m12A r !
.
3
k ps p
12R 2r e 22 ~ b r m1A r !
~8f!
and
Expressions for H 1 ,H 2 and G 3 ,G 4 are given in the Appendix. From the structure of the PPE output voltage of Eq.
~7!, it is obvious that the overall output signal is the result of
the complex ~vectorial! superposition of the thermal-wave
fields within the PVDF detector generated by three sources:
~1! direct transmission of the incident light passing through
the sample and the reference @the H 1 G 1 and H 2 G 2 terms in
the numerator of Eq. ~7!#; ~2! thermal-wave confinement
within the cavity formed by the sample and the reference
@ W 21 and V 34 terms in the numerator of Eq. ~7!#; and ~3!
nonradiative ~optical-to-thermal! conversion processes of the
sample and the reference following optical absorption. These
occur in bulk and at the surface @H 1 G 3 and H 2 G 4 terms in
~7!
the numerator of Eq. ~7!#. Theoretical simulation shows that
the contribution from the third source is much smaller
~;3–4 orders of magnitude less! than that from the first and
the second sources for highly transparent materials.19 Therefore, the third source has been neglected in the quantitative
analyses of our data. If A s 50 and A r 50, the PPE output of
Eq. ~7! reduces to the monolithic expression, Eq. ~5!, of Ref.
18, as expected.
It can be seen that the overall output PPE signal of Eq.
~7! is affected by the surface absorptances as well as by the
bulk absorption coefficients of the sample and the reference.
By using appropriate combinations of pairs of samples, in
which only one parameter ~surface absorptance, bulk absorption or thickness!, varies between a sample-reference pair,
one may precisely measure the variation of this parameter
without requiring accurate information about the values of
other parameters. This advantage is attributed to the nature
of the destructive interferometric measurement, in which
contributions from the same optical/thermal property are coherently canceled within the PVDF detector, while only the
signal difference due to small variations in one parameter
appears. In contrast, the optical parameters of the sample are
always inherently coupled together in the single-ended PPE
measurements. Therefore, in the single-ended configuration,
one must know all other parameters in order to measure one
of them, and the measurement precision is limited by the
precision associated with this knowledge. Instrumentally, the
coherent noise cancellation within the single detector in the
interferometric PPE method is superior to conventional differential optical measurements,22 in which electronic noise
components owing to the employment of two photonic de-
Downloaded 18 Jul 2008 to 128.100.49.17. Redistribution subject to AIP license or copyright; see http://rsi.aip.org/rsi/copyright.jsp
3118
Rev. Sci. Instrum., Vol. 70, No. 7, July 1999
C. Wang and A. Mandelis
nm! to that at the emission peak ~820 nm!.3 A high FOM
implies a high-quality bulk. Laser-rod grade materials are
characterized by FOM ;1000.
1. Samples with identical bulk but different surface
quality
FIG. 2. Experimental setup for PPE thermal-wave interferometry. s: sample;
p: PVDF detector; r: reference.
tectors and an electronic signal ratioing amplifier, add, rather
than cancel out.
III. EXPERIMENTAL AND RESULTS
Two samples with the same FOM ~5800! were measured. One sample did not have optimal ~henceforth, ‘‘nonoptimal’’! surface polish, and thickness50.0810 cm; the
other sample had laser-rod-grade ~‘‘good’’! polish and
thickness50.0771 cm. Both samples had been cut from the
same Ti sapphire rod and were subjected to the same crystal
growth procedure, including post-growth anneal. The nonoptimal crystal was polished with a diamond paste containing 5
mm size particulates. The ‘‘good’’ crystal was polished with
the best available surface mechanical polish, using a diamond paste with 1 mm size particulates, followed by a further mechanical polish using 0.25 mm size diamond particulates.
A. Experimental setup
The PPE interferometric setup has been described
previously.17–19 A schematic is shown in Fig. 2. At the heart
of this interferometer lie two thermal-wave cavities formed
by sample-pyroelectric and pyroelectric-reference compartments. The PVDF thin film detector, 52 mm thick and 2 cm
in diameter, was installed on an aluminum-base bearing a
hole. The PVDF element acts as a thermal-wave signal transducer and as a wall for front and back thermal-wave cavities.
Both sample and reference are mounted on a threedimensional ~3D! angularly and linearly adjustable micrometer stage of 10 mm resolution in linear motion and 0.1° in
angular rotation. The relative intensities of the front and back
incident beams, which are split off of a He–Ne laser ~l
5632.8 nm, P'5 mW!, are adjusted by a linear intensity
attenuator, and the phase shift between the two beams is
precisely controlled by a mechanical chopper ~EG&G Model
192!, also fixed on a micrometer stage. The experimental
data are collected by a PC via a lock-in amplifier ~EG&G
Model 5210!. Several pairs of Ti-sapphire crystals were used
in the role of sample and reference in our experiments. For
each pair of crystals, only one parameter was different: either
the surface absorptance, or the bulk absorption coefficient, or
the thickness. The reference was fixed in either the OT
~purely optical mode! or the PPE mode,1 depending on the
difference between the sample and the reference. The
sample-PVDF cavity length was scanned. By fitting the scanning curves ~amplitude and phase! to Eq. ~7! for each pair of
sample-reference combination, measurements of the desired
parameters were obtained.
B. Sample description and preparation
The Ti-sapphire crystals used in this work were grown
by the Czochralski pulling technique at Union Carbide,
Washougal, WA. Four pairs of crystals with different FOM,
different surface polish treatments, and different thicknesses
were measured. The FOM in Ti31:Al2O3 is defined as the
ratio of the absorption coefficient at the absorption peak ~490
2. Samples with different bulk but identical surface
quality
Two samples ~FOM540, thickness52.017 cm; and
FOM5800, thickness52.013 cm! were used. They were
subjected to the same nonoptimal surface polishing process
as described above. The difference in bulk optical quality
was due to difference in growth processes. The two crystals
were grown in an identical manner using the Czochralski
technique. Then, the FOM5800 sample was subjected to
further annealing, thereby removing bulk optical defects
present in the unannealed crystal.2
3. Samples with identical bulk and surface quality but
different thickness
Two pairs of samples were used. The first pair included
samples of thicknesses 0.0810 and 0.7929 cm, of the same
FOM ~5800!, and the same nonoptimal surface polish. The
second pair consisted of samples of thicknesses 0.0771 and
1.0677 cm, of bulk FOM5800, and the same ‘‘good’’ surface polish, as described above.
C. Results
Each pair of the Ti:sapphire samples was measured, with
one crystal acting as the sample and the other crystal acting
as the reference. The general experimental procedure for all
the measurements was as follows: The relative intensities
and the phase shift between the two incident beams were
adjusted, such that the demodulated lock-in output equaled
zero before the sample and the reference were put into place.
This procedure makes the PPE system operate in the fully
destructive interferometric mode when both the sample and
the reference are absent, i.e., I 1 5I 2 , and D w 5180. The
sample and the reference were then inserted into the optical
path. For each measurement, the cavity length between the
sample and the PVDF detector was scanned between the
thermally uncoupled and strongly coupled limits.
Downloaded 18 Jul 2008 to 128.100.49.17. Redistribution subject to AIP license or copyright; see http://rsi.aip.org/rsi/copyright.jsp
Rev. Sci. Instrum., Vol. 70, No. 7, July 1999
FIG. 3. Experimental and theoretical ~fitted! results of ~a! the amplitude and
~b! phase for a pair of Ti:sapphire crystals with identical bulk optical properties ~FOM!, but different surface polish. Solid squares: experimental; solid
lines: theoretical fits. Insets: experimental and theoretical fit results for the
same pair of samples, when the reference was placed farther away from the
PVDF sensor ~relatively shallow PPE mode!.
The reference was fixed in either the thermally uncoupled or
fully coupled mode, depending on the particular sample pair
combination.
To study samples with identical bulk quality but different surface preparation, the crystal with good surface polish
~thickness l50.0771 cm! was used as the sample, and the
crystal with nonoptimal polish ~thickness m50.0810 cm!
was used as the reference. Known parameters of the sample
and the reference are:2 a s 5 a r 50.106 cm2/s, k s 5k r
50.33 W/cm K, R s 5R r 50.07, and b s 5 b r 50.04 cm21.
The modulation frequency was f 526.5 Hz. Figure 3 shows
the scanning curves of the PPE ~a! amplitude and ~b! the
phase when the reference was placed in the deep PPE mode
~very close to the PVDF detector!. By fitting the experimental data of the amplitude channel to the theoretical formula
Eq. ~7!, the difference in surface absorptance between the
sample and the reference (DA5A r 2A s ) was found to be
0.011360.0002. The cavity length, L 1 , between the PVDF
film and the reference was 0.21360.002 mm, the average of
three measurements. The PPE phase was then calculated using DA and L 1 obtained from the amplitude fit to ascertain
consistency and validity. The multiparameter fit allows for
four parameters to be calculated uniquely from Eq. ~7!:
Bulk and surface optical absorptions
3119
S( v ); DA; absolute value of L 1 ; and DL, a correction for
zero separation between the sample and the surface of the
detector, which was used to yield absolute value of L, i.e., L
is equal to the experimental position plus the offset value
DL.
To check the sensitivity and to validate the measurement, the reference was subsequently moved farther away
from the detector by 0.05 mm, i.e., it was set in a relatively
shallow PPE mode. The measurement curves are shown in
the insets of Figs. 3~a! and 3~b! for the amplitude and the
phase, respectively. The fitted values for DA and L 1 at this
position were 0.011660.0005 and 0.26760.003 mm, respectively. It was therefore concluded that both measurements gave very consistent values, the relative error between
the two measurements being only 2.7% for DA. The fitted
value difference for the cavity length L 1 between the two
measurements is 0.054 mm, in excellent agreement with the
actual scanned distance of 0.05 mm. In view of the experimental readout error from the 10 mm resolution micrometer
stage, the reliability of the theoretical fits was judged to be
excellent. As regards the SNR and reproducibility, it is noted
that the deep PPE mode ~short L 1 ! is better than the shallow
PPE mode ~longer L 1 !, due to the higher signal levels resulting from better thermal-wave confinement ~higher thermalpower density! inside the cavity in the former configuration.
In the foregoing fitting process, the bulk absorption coefficient and the surface reflectivity of the sample ~which
was the same as that of the reference!, were set to be 0.04
cm21 and 0.07, respectively. However, it was found that the
fitted DA value changes from 0.0113 to 0.0099, only a ;4%
variation, when the value of the bulk absorption coefficients
b s (5 b r ) was varied from 1028 to 0.4 cm21, in the case of
the two samples with different thickness ~l50.0771 cm, m
50.0810 cm!. Further simulations showed that the fitted DA
value is substantially independent of the value of the bulk
absorption coefficient b s (5 b r ), if the thicknesses of the
sample and the reference are identical. This is an important
feature of the interferometric PPE technique. It implies that
for the purpose of measuring the difference of surface absorptances between two crystals of the same ~or nearly
equal! thickness and bulk optical quality, it is not necessary
to know the exact value of the bulk absorption coefficient. A
similar observation was made regarding surface reflectivity
(R s 5R r ): the fitted value of DA changes from 0.0118 to
0.0111 ~;5.9%! when the values of R s (5R r ) are varied
between 0.02 and 0.08. This can be attributed to the thermalwave destructive interferometric effect, due to which equal
variations of the same parameter in both materials are cancelled to the large extent. In practice, it would be helpful to
choose two thin samples in order to minimize the effect of
the bulk absorption coefficient to the overall absorptance and
to increase the tolerance of the thickness difference between
the sample and the reference. Thus, the dependence of the
fitted value DA on the initial value of the bulk absorption
coefficient ( b s 5 b r ) becomes trivial.
To study samples with different bulk optical properties,
but identical surface preparation, one Ti sapphire crystal
(FOM5800) of thickness l52.013 cm was used as the
sample. The other crystal (FOM540) of thickness m
Downloaded 18 Jul 2008 to 128.100.49.17. Redistribution subject to AIP license or copyright; see http://rsi.aip.org/rsi/copyright.jsp
3120
Rev. Sci. Instrum., Vol. 70, No. 7, July 1999
C. Wang and A. Mandelis
FIG. 4. Experimental and theoretical ~fitted! results of ~a! the amplitude and
~b! phase for a pair of Ti:sapphire crystals with identical surface polishes,
but different bulk optical properties ~FOM!.
52.017 cm was used as the reference. The modulation frequency was 10 Hz. Figures 4~a! and 4~b! show the experimental results and the theoretical fits, when the reference was
placed in the thermally uncoupled ~OT! mode. In practice,
this occurs when the reference cavity length L 1 , Fig. 1, is
greater than 3 mm. The use of the large-cavity-length OT
mode for the reference cavity is dictated by the appearance
of the sharp minimum in the signal amplitude, when the total
absorptances of sample and reference are close to each other,
e.g., Fig. 5~a!. The line shape and position of the minimum
are used as a sensitive measure for material property calculations. This minimum does not appear when the reference
cavity length decreases, so that thermal confinement affects
the signal ~PPE mode!.
In Fig. 4, the experimental amplitude data were fitted to
the theoretical formula Eq. ~7! using the following parameters: A s 5A r 50.05 and R s 5R r 50.07. The other parameters
are the same as those used in Fig. 3. The difference of the
bulk absorption coefficient between the FOM540 and the
FOM5800 Ti: sapphire crystals was found to be: D b 5 b r
2 b s 50.054560.0006 cm21, the average of three measurements. The cavity length L 1 was found to be greater than 8
mm for all three measurements. This best-fit result is consistent with the experimental arrangement, in which the reference was actually placed in OT mode. Both amplitude and
phase fits to the data are excellent, as shown in Figs. 4~a! and
4~b!. The computational procedure conducted by using different values of surface absorptance A s (5A r ) and surface
reflectivity R s (5R r ) found that the change in the fitted value
of Db was only 2.1% ~from 0.0552 to 0.0540 cm21! and
FIG. 5. Experimental and theoretical ~fitted! results of ~a! the amplitude and
~b! phase for a pair of Ti:sapphire crystals of identical bulk optical properties ~FOM! and surface ~‘‘nonoptimal’’! polish, but different thicknesses
(l50.0810 cm; m50.7929 cm!. Solid squares with error bar: experimental
data; solid lines: theoretical fits. Insets: experimental and theoretical ~fitted!
curves for another pair of crystals with identical bulk optical properties
~FOM! and surface ~‘‘good’’! polish, but different thicknesses ~l
50.0771 cm, m51.0677 cm!.
2.2% ~from 0.0549 to 0.0537 cm21!, respectively, when A s
was changed from 1025 to 1021 and R s was changed from
0.02 to 0.13. Once again, it is shown that the measurement of
bulk Db does not depend strongly on the absolute values of
surface absorptance and reflectivity over wide ranges of A s
and R s . From a different viewpoint, the difference in bulk
absorption coefficients between a sample and a reference
crystal can be precisely measured without accurate knowledge of the surface absorptance and reflectivity values. It
should be noted that the absence of the minimum ~‘‘dip’’! in
Fig. 4~a!, unlike Fig. 3~a!, is caused by the large difference in
transmitted radiation past the sample and the reference crystals, owing to their very different bulk absorption coefficients. Therefore, the interference between the two thermalwave fields cannot become destructive ~or nearly so!.
Furthermore, the thermal-wave power confinement effect between the sample and the PVDF detector at short cavity
lengths is not sufficiently strong to compensate for such a
large signal difference caused by the direct optical transmissions on both sides of the detector.
Finally, two pairs of crystals were measured, each pair
with identical bulk optical properties (FOM5800) and surface preparation ~polish!, but with different thicknesses. One
Downloaded 18 Jul 2008 to 128.100.49.17. Redistribution subject to AIP license or copyright; see http://rsi.aip.org/rsi/copyright.jsp
Rev. Sci. Instrum., Vol. 70, No. 7, July 1999
pair ~thicknesses l50.0810 cm and m50.7929 cm! had the
same nonoptimal polish; the other pair ~thicknesses l
50.0771 cm and m51.0677 cm! had the same good polish.
For each pair, the thicker sample was employed as the reference and was placed in the OT mode. The modulation frequency for both cases was f 510 Hz. The measurement
scans ~amplitudes and phases! are shown in Figs. 5~a! and
5~b!. By fitting the experimental amplitude of one pair to the
Eq. ~7!, the absolute bulk absorption coefficient of the two
crystals was found to be: b s (5 b r )50.061360.0004 cm21,
the average of three independent measurements. The multiparameter fits yielded the cavity length L 1 .3 mm for all
three measurements. The values of surface absorptance and
reflectivity were assumed to be: A s 5A r 50.06, and R s 5R r
50.07, respectively. Similarly, the absolute bulk absorption
coefficients for the other pair were found to be: b s (5 b r )
50.060460.0006 cm21, and L 1 .8 mm. The values obtained for the cavity length L 1 are consistent with the experimentally selected OT mode for the reference. It is interesting
to note that the calculated absolute bulk absorption coefficients of the two pairs of crystals were in excellent agreement, the relative error between the two measurements being
only about 1.5%.
Similar to earlier findings, the fitted results for b s
(5 b r ) do not depend strongly on the values of A s (5A r ) and
R s (5R r ): the b s value changed from 0.0619 to 0.0602 cm21
~;2.7%! with A s variations between 631024 and 1021 ; b s
changed from 0.0617 to 0.0603 cm21 ~;2.3%! with R s
variations between 0.01 and 0.12. Therefore, again, it may be
concluded that the ~common! absolute bulk absorption coefficient of a crystal pair can be obtained precisely by means of
PPE destructive interferometry, using two samples with different thicknesses. In the present case, the total thermal-wave
fields at both sides of the PVDF detector for both pairs of
samples were just close enough to produce a minimum in the
amplitude of Fig. 5~a! and its inset when the reference is in
the OT mode. This would not be the case when the reference
is placed in the PPE mode for these two pairs of samples.
Bulk and surface optical absorptions
3121
FIG. 6. Theoretical effect of reference-cavity length on the signal output of
a PPE thermal-wave interferometer. The modulation frequency is f
526.5 Hz. The assumed values of the material parameters are the same as
those used in Fig. 3 with DA50.0113.
IV. DISCUSSION
PPE signal minimum. On the other hand, for a pair of
samples with relatively small differences in optical absorptance, a thermally thin reference cavity length ~the PPE
mode! is preferred. Figure 6 shows theoretical curves justifying this statement at f 526.5 Hz. The parameters used in
that simulation are the same as those in Fig. 3 with DA
50.0113. From this simulation it can be seen that the deepest PPE mode ~shortest reference cavity length; L 1
50.1 mm! gives the sharpest interferometric dip in the amplitude channel and the largest dynamic range in the phase
channel. Therefore, higher quality measurements can be expected by using strongly thermally thin reference cavities,
from the point of view of the superior signal-to-noise ratio
and the resulting precision of the fitting process, as discussed
in relation to Fig. 3.
A. Effect of PVDF—reference-solid separation
„reference cavity length…
B. Effect of the operating thermal-wave frequency
The experimental results, shown in Figs. 3, 4, and 5,
indicate that the PPE signal can be affected by the operating
~OT or PPE! mode of the reference. The selection of the
reference-PVDF distance in terms of thermal thickness of the
intracavity gas ~air! layer,1 should be guided by the goal to
minimize the intensity differences between the thermal-wave
fields caused by the directly transmitted light on both sides
of the PVDF detector, so as to improve the quality of the
measurements. For a pair of samples with relatively large
difference in optical absorptances, caused either by surface,
bulk, or geometrical parameters, the OT mode for the reference is optimal. This is so, because the thermally thick gas
layer in the reference cavity does not introduce a thermal
component to the reference signal to offset the thermal confinement in the sample cavity. This allows the latter to exhibit its maximum effect in the form of the observed overall
To optimize measurements, the characteristics of the
PPE output signals under various modulation frequencies are
calculated and shown in Fig. 7 using the parameters of Fig.
3. The cavity length L 1 is fixed at 0.2 mm. It can be seen
from Fig. 7 that the signal dynamic range in both amplitude
and phase channels at all frequencies considered in the simulation is comparable, however, sharper interferometric amplitude minima appear at lower frequencies. As a result, a
low operating frequency should be employed, in order to
produce a sharper interferometric ‘‘dip’’ and a higherprecision measurement. There is a limit in precision, however, since the lower modulation frequency will introduce a
larger 1/f electronic noise leading to a degraded SNR. Therefore, in practice, the modulation frequency is usually chosen
between 10 and 30 Hz. This frequency characteristic is also
experimentally demonstrated in Figs. 8~a! and 8~b! at two
Downloaded 18 Jul 2008 to 128.100.49.17. Redistribution subject to AIP license or copyright; see http://rsi.aip.org/rsi/copyright.jsp
3122
Rev. Sci. Instrum., Vol. 70, No. 7, July 1999
C. Wang and A. Mandelis
FIG. 7. Theoretical effect of modulation frequency on the PPE signal output
of a thermal-wave interferometer. The reference is fixed at L 1 50.2 mm.
The assumed values of the material parameters are the same as those used in
Fig. 3 with DA50.0113.
FIG. 9. Solid squares: theoretical curve fitted to the PPE experimental data
of Fig. 3, DA50.0113; and open circles: using the same instrumental constant, and assuming DA5531024 . Other parameters of the material are the
same as those in Fig. 3.
different frequencies ~f 510 and 26.5 Hz! using the same
samples and experimental conditions as that in Fig. 5.
position of the minimum should be zero when the sample
and the reference are identical ~i.e., DA50, D b 50, and l
5m!. If one of these optical parameters ~e.g., surface absorptance! changes, DAÞ0, the output signal at the dip is no
longer equal to zero. Based on this nonzero dip output, we
can obtain the small difference DA by means of a theoretical
fit to the signal. The smallest measurable nonzero dip output
will determine the maximum resolution of the measurement.
Figure 9 shows the theoretical curve ~solid squares! fitted to
the experimental data in Fig. 3, from which the difference of
the surface absorptance between the sample and the reference (DA50.0113) and an instrumental constant
S(26.5 Hz)50.65126 a.u., were obtained ~see also Sec.
III C!. By using this instrumental constant and assuming a
very small difference in the surface absorptance, DA55
31024 , the other theoretical curve ~Fig. 9, open circles! was
calculated. From this theoretical simulation, it was found that
the output at the interferometric dip is approx. 0.5 mV. This
level of PPE signal output is actually at the system noise
limit in our present setup. Therefore, it is believed that the
maximum resolution for measuring differential surface absorptance can reach the level of ;1024 .
A similar procedure can be followed to estimate the
maximum resolution of the technique for measuring bulk
absorption coefficient differences, Db. In this case, the instrumental constant, S(10 Hz)50.31091 a.u. was obtained
from the best-fit results of Fig. 4. With the help of this instrumental constant, the theoretical curve in Fig. 10 was calculated using D b 5231024 cm21, so that the signal output
would again be 0.5 mV, which is the system noise limit of
our present setup. For maximum resolution, the reference
was assumed to be fixed at L 1 50.2 mm ~PPE mode!. Therefore, ;1024 cm21 of differential bulk absorption can be considered as the maximum resolution for the pair of samples of
C. Measurement resolution
The
maximum
achievable
differential-surfaceabsorptance, or bulk-absorption-coefficient resolution between the sample and the reference depends on the noise
level of the output signal at the position of the interferometric minimum ~dip!. Theoretically, the output signal at the
FIG. 8. Experimental demonstration of the effect of two different frequencies on the PPE interferometric signal using the same samples and experimental conditions as that in Fig. 5.
Downloaded 18 Jul 2008 to 128.100.49.17. Redistribution subject to AIP license or copyright; see http://rsi.aip.org/rsi/copyright.jsp
Rev. Sci. Instrum., Vol. 70, No. 7, July 1999
Bulk and surface optical absorptions
3123
to be useful in quality control of both bulk crystal growth
and surface treatment of optical and laser crystals, thus impacting the performance of solid-state lasers.
ACKNOWLEDGMENT
The support of the Natural Sciences and Engineering
Research Council of Canada ~NSERC! through a Research
Grant is gratefully acknowledged.
APPENDIX: DEFINITIONS OF EXPRESSIONS
N 1 [11
R s R p ~ 12R s !~ 11R s e 2A s ! e 22 ~ b s l1A s !
12R 2s e 22 ~ b s l1A s !
F
N 2 [e 2A s R s 1
N 1r [11
FIG. 10. Theoretical PPE signal output using the same instrumental constant
as that in Fig. 4 and assuming D b 5231024 cm21. Other parameters of the
material are the same as those in Fig. 3.
R p ~ 12R s !~ 11R s e 2A s !
12R 2s e 22 ~ b s l1A s !
G
~A2!
,
R r R p ~ 12R r !~ 11R r e 2A r ! e 22 ~ b r m1A r !
12R 2r e 22 ~ b r m1A r !
F
N 2r [e 2A r R r 1
R p ~ 12R r !~ 11R r e 2A r !
12R 2r e 22 ~ b r m1A r !
~A1!
,
G
, ~A3!
~A4!
.
Also
thickness ;2 cm. Nevertheless, the maximum possible resolution can easily reach 1025 – 1026 cm21 by using pairs of
longer samples, say >5 cm, or higher laser fluence.
Using the same estimation procedure for technique resolution regarding the absolute bulk absorption coefficient, if
two transparent samples with thicknesses differing by one
order of magnitude are used, the resolution can be estimated
to be 1025 – 1026 cm21.
Based on the extension of our earlier purely thermalwave photopyroelectric interferometric theory, the differential surface absorptance and differential bulk absorption coefficient, as well as the absolute bulk absorption coefficient
of Ti:sapphire crystals have been separately and precisely
measured using appropriate sample combinations. Owing to
the destructive PPE interferometric effect, and unlike the
conventional single-ended PPE technique, this interferometric method does not require precise knowledge of the remaining sample optical parameters to produce a highprecision measurement of one of optical parameters. The
resolution limits for small differences in surface absorptance
and bulk absorption coefficient were estimated to be ;1024
and 1025 – 1026 cm21, respectively, with the current instrumental setup ~the laser power is only ;5 mW!. The resolution for low absolute bulk absorption coefficient was estimated to be 1025 – 1026 cm21. This is similar or even
superior to other established, yet experimentally more involved photothermal techniques,5,8 such as the photothermal
deflection technique ~PTD!, in which a high-power pumping
laser ~;5 W! and a probing laser of a good Gaussian beam
spot have to be employed and delicately adjusted with respect to the relative position of the position-sensing photodetector in order to obtain comparable resolution. In addition, a calibration process has to be conducted in PTD.5,8
This new thermal-wave interferometric technique is expected
I t ~ 0 ! 5I 1
I t ~ l ! 5I 1
~ 12R s ! e 2A s
12R 2s e 22 ~ b s l1A s !
~ N 1 1N 2 e 22 b s l ! ,
~ 12R s ! e 2 ~ b s l1A s !
12R 2s e 22 ~ b s l1A s !
I tr ~ 0 ! 5I 2 e jD w
~A5!
~A6!
~ N 1 1N 2 ! ,
~ 12R r ! e 2A r
12R 2r e 22 ~ b r m1A r !
~ N 1r 1N 2r e 22 b r m ! ,
~A7!
I tr ~ m ! 5I 2 e jD w
I 1 ~ l ! 5I 1
~ 12R r ! e
2 ~ b r m1A r !
12R 2r e 22 ~ b r m1A r !
~ 12R s ! 2 e 2 ~ b s l12A s !
12R 2s e 22 ~ b s l1A s !
I 2 ~ m ! 5I 2 e j w
~A8!
~ N 1r 1N 2r ! ,
~A9!
,
~ 12R r ! 2 e 2 ~ b r m12A r !
12R 2r e 22 ~ b r m1A r !
~A10!
.
The expressions for the constants H 1 , H 2 , G 3 , and G 4 in
Eq. ~7! are
H 1 5 ~ e s p d 21 !~ 11 g gp V 34e 22 s g L 1 !
1 ~ 12e 2 s p d !~ g gp 1V 34e 22 s g L 1 ! ,
~A11!
H 2 5 ~ e s p d 21 !~ 11 g gp W 21e 22 s g L !
1 ~ 12e 2 s p d !~ g gp 1W 21e 22 s g L ! ,
G 35
2Q 1 e
2ssl
2Q 2 ~ 11b gs ! 1Q 3 ~ 12b gs ! e
~A12!
22 s s l
~ 11b gs ! 2 ~ 12 g 2gs e 22 s s l !
,
~A13!
with
Q 1 5E s @ b gs ~ N 1 1N 2 e 22 b s l ! 1r s ~ N 1 2N 2 e 22 b s l !#
1 h ~s0 ! A s I t ~ 0 ! / ~ k s s s ! ,
~A14!
Downloaded 18 Jul 2008 to 128.100.49.17. Redistribution subject to AIP license or copyright; see http://rsi.aip.org/rsi/copyright.jsp
3124
Rev. Sci. Instrum., Vol. 70, No. 7, July 1999
C. Wang and A. Mandelis
Q 2 5E s e 2 b s l @ N 1 1N 2 1r s ~ N 1 2N 2 !#
1
2 h ~s0 ! A s I t ~ l ! / ~ k s s s ! ,
~A15!
Q 3 5E s e 2 b s l @ N 1 1N 2 2r s ~ N 1 2N 2 !#
1 h ~s0 ! A s I t ~ l ! / ~ k s s s ! ,
E s5
I 1 h bs b s
•
~ 12R s ! e 2A s
2k s ~ b 2s 2 s 2s ! 12R 2s e 22 ~ b s l1A s !
~A16!
~A17!
,
r s5 b s / s s ,
~A18!
and
G 45
2 P 1 e 2 s r m 1 P 2 ~ 11b gr ! 2 P 3 ~ 12b gr ! e 22 s r m
~ 11b gr ! 2 ~ 12 g 2gr e 22 s r m !
,
~A19!
with
P 1 5E r @ b gr ~ N 1r 1N 2r e 22 b r m ! 1r r ~ N 1r 2N 2r e 22 b r m !#
1 h ~r0 ! A r I tr ~ 0 ! / ~ k r s r ! ,
~A20!
P 2 52E r e 2 b r m @ N 1r 1N 2r 1r r ~ N 1r 2N 2r !#
1 h ~r0 ! A r I tr ~ m ! / ~ k r s r ! ,
~A21!
P 3 52E r e 2 b r m @ N 1r 1N 2r 2r r ~ N 1r 2N 2r !#
2 h ~r0 ! A r I tr ~ m ! / ~ k r s r ! ,
E r5
I 2 e jD w h br b r
•
~ 12R r ! e 2A r
2k r ~ b 2r 2 s 2r ! 12R 2r e 22 ~ b r m1A r !
r r5 b r / s r .
A. Mandelis and M. M. Zver, J. Appl. Phys. 57, 4421 ~1985!.
A. Mandelis, J. Vanniasinkam, S. Buddhudu, and M. Kokta, Phys. Rev. B
48, 6808 ~1993!.
3
J. Vanniasinkam, A. Mandelis, S. Buddhudu, and M. Kokta, J. Appl. Phys.
75, 8090 ~1994!.
4
J. Vanniasinkam, A. Mandelis, M. Munidasa, and M. Kokta, J. Opt. Soc.
Am. B 15, 1647 ~1998!.
5
A. P. Kubyshkin, M. P. Matrosov, and A. A. Karabutov, Opt. Eng. ~Bellingham! 33, 3214 ~1994!.
6
D. Dadarlat, M. Chirtoc, R. M. Candea, and I. Bratu, Infrared Phys. 24,
469 ~1984!.
7
A. C. Boccara, D. Fournier, W. Jackson, and N. H. Amer, Opt. Lett. 5,
377 ~1980!.
8
Y.-X. Nie and L. Bertrand, J. Appl. Phys. 65, 438 ~1989!.
9
A. Rosencwaig and T. W. Hidley, Appl. Opt. 20, 606 ~1981!.
10
J. Vanniasinkam, M. Munidasa, A. Othonos, M. Kokta, and A. Mandelis,
IEEE J. Quantum Electron. 33, 2301 ~1997!.
11
J. Shen, K. Fjelsted, J. Vanniasinkam, and A. Mandelis, Opt. Mater. 4,
823 ~1995!.
12
S. Buddhudu, A. Mandelis, B. Joseph, and K. Fjelsted, Opt. Mater. 3, 115
~1994!.
13
E. Welsch, in Handbook of Optical Properties, Volume I, Thin Film for
Optical Coatings, edited by R. E. Hummel and K. H. Guenther ~CRC,
Boca Raton, FL, 1995!, p. 243.
14
A. Mandelis, R. E. Wagner, K. Ghandi, R. Baltman, and P. Dao, Phys.
Rev. B 39, 5254 ~1989-I!.
15
C. Christofides, K. Ghandi, and A. Mandelis, Meas. Sci. Technol. 1, 1363
~1990!.
16
C. Christofides, A. Mandelis, K. Ghandi, and R. E. Wagner, Rev. Sci.
Instrum. 61, 2360 ~1990!.
17
C.-H. Wang and A. Mandelis, Proceedings of the X International Conference on Photoacoustic Photothermal Phenomena, edited by F. Scudieri
~AIP, New York, 1998!, p. 138.
18
C.-H. Wang and A. Mandelis, J. Appl. Phys. 85, 8366 ~1999!.
19
C.-H. Wang and A. Mandelis, Rev. Sci. Instrum. ~in press!.
20
M. R. Kokta, in Tunable Solid-State Lasers II, edited by A. B. Budgor, L.
Esterowitz, and L. G. De Shazer ~Springer, New York, 1986!, p. 89.
21
J. F. Pinto, L. Esterowitz, G. H. Rosenblatt, M. Kokta, and D. Peressini,
IEEE J. Quantum Electron. 30, 2612 ~1994!.
22
P. H. Klein, Physical Properties of Optical Materials ~SPIE, Bellingham,
WA, 1979!, Vol. 204, p. 77.
2
~A22!
,
~A23!
~A24!
Downloaded 18 Jul 2008 to 128.100.49.17. Redistribution subject to AIP license or copyright; see http://rsi.aip.org/rsi/copyright.jsp