Zoll Aed Plus Manual
Zoll Aed Plus Manual
Zoll Aed Plus Manual
Warnings
• Use the ZOLL AED Plus unit only as described in this manual. Improper use of the device can cause
death or injury.
• DO NOT use or place the ZOLL AED Plus unit in service until you have read the ZOLL AED Plus
Operator’s and Administrator’s Guides.
• DO NOT use or place the ZOLL AED Plus unit in service if the unit’s status indicator window
(located on the left side of the handle) displays a red “X”.
• DO NOT use or place the ZOLL AED Plus unit in service if the unit emits a beeping tone.
• Connect the electrode cable to the ZOLL AED Plus unit after installing batteries.
• Keep the electrode cable connected to the ZOLL AED Plus unit at all times.
• This device should only be used by properly trained individuals.
• Only use electrodes labeled “Infant/Child” on children less than 8 years old or weighing less than
55 lbs (25 kg). Use CPR-D padz if patient is older than 8 years or weighs more than 55 lbs (25 kg).
• Always stand clear of patient when delivering treatment. Defibrillation energy delivered to the
patient may be conducted through the patient’s body and cause a lethal shock to those touching the
patient.
• DO NOT TOUCH the electrode surfaces, the patient, or any conductive material touching the patient
during ECG analysis or defibrillation.
• Move patient away from electrically conductive surfaces prior to use of equipment.
• DO NOT use this unit near or within puddles of water.
• Keep the patient as motionless as possible during ECG analysis.
• DO NOT use this unit near flammable agents, such as gasoline, oxygen-rich atmospheres or
flammable anesthetics.
• Avoid radio frequency interference from high-power sources that might cause the defibrillator to
interpret cardiac rhythms incorrectly by turning off cell phones and 2 way-radios.
• Disconnect non-defibrillation protected electronic devices or equipment from patient before
defibrillation.
• Dry patient’s chest, if wet, before attaching electrodes.
• Apply freshly opened and undamaged electrodes, within the electrode expiration date, to clean and
dry skin to minimize burning.
• DO NOT place the electrodes directly over the patient’s implanted pacemaker. Pacemaker stimuli
may degrade the accuracy of ECG rhythm analyses or the pacemaker may be damaged by
defibrillator discharges.
• Check labeling inside the ZOLL AED Plus cover before using the cover as a Passive Airway Support
System (PASS) device. Ensure it is intended for this use.
• DO NOT use Passive Airway Support System (PASS) if there is a suspected head or neck injury.
Place patient on a firm surface before performing CPR.
• DO NOT recharge, disassemble, or dispose of batteries in fire. Batteries may explode, if mistreated.
• Do not use or stack the ZOLL AED Plus unit with other equipment. If the unit is used or stacked with
other equipment, verify proper operation prior to use.
If any of the events described above occur, please contact ZOLL Medical Corporation in writing with the
following information:
1. Originator’s organization - Company Name, Address, Contact Name and Contact Phone Number.
2. Part Number/Model Number and Serial Number.
3. Disposition of Device (e.g. received, lost, stolen destroyed, distributed to another organization).
4. New Location and/or Organization (if different from #1 above) - Company Name, Address, Contact
Name and Contact Phone number.
5. Date change took effect.
Unpacking
• Carefully inspect each packing container for damage.
• Examine the unit for any signs of damage that may have occurred during shipping.
• If the contents are incomplete or damaged or if the unit fails to pass its self test as indicated by a Red
“X” in the status indicator window after battery installation, contact ZOLL Medical Corporation’s
Technical Service Department.
• Review the shipping list to insure that all items ordered were received.
Conventions
Throughout this document, voice prompts are indicated by capital italicized letters, such as
CALL FOR HELP.
WARNING! Warning statements describe conditions or actions that can result in personal injury or
death.
CAUTION! Caution statements describe conditions or actions that can result in damage to the
unit.
Class II equipment
DANGEROUS VOLTAGE
Push button
Manufacturer
Serial Number
Catalogue number
WARNING! The CPR function is intended for use on adult patients only; do not use on patients
under 8 years of age.
The metronome function is disabled during periods when CPR should not be performed (for example,
during ECG analyses and defibrillation shock sequences). During periods when CPR may be indicated,
the metronome begins issuing audible beeps following detection of the rescuer’s first few
compressions. The beeps continue automatically (at rates described below) until a few seconds after
chest compressions are halted by the rescuer or until the recommended “CPR period” ends (1 minute
for AHA and ERC protocols). If the rescuer ceases chest compressions during the CPR period,
metronome beeps stop within a few seconds after compressions are halted. Audible beeps resume
during the CPR period following any re-initiation of CPR compressions. If no CPR compressions are
detected during “CPR periods”, the ZOLL AED Plus periodically re-issues the CONTINUE CPR
prompt.
The rate of beeps issued by the ZOLL AED Plus metronome function adapts to the rescuer’s actual
chest compression rate. The metronome will beep at 100 CPM when chest compressions are delivered
at greater than 80 compressions per minute (CPM). Should the rescuer fail to deliver compressions at
80 CPM or greater, the metronome will beep at a rate that is approximately 15 CPM higher than the
rescuer’s actual rate. This increased metronome rate is intended to encourage the rescuer to increase
his/her chest compression rate until the recommended 100 CPM rate is achieved. The metronome beeps
at a minimum rate of 60 CPM in cases where the rescuer’s compression rate is substantially below
60 CPM.
During CPR, the ZOLL AED Plus may issue one or more audible prompts based on the depth of chest
compressions detected. When the CPR monitoring system determines that compression depth is
consistently less than 1½ inches, a PUSH HARDER prompt will be issued. A GOOD COMPRESSION
prompt will be issued if the rescuer responds by increasing compression depth to 1½ inches or more.
Status
Indicator
On/Off
Speaker (bottom)
Shock Button
Control/Indicator Description
Indicator lights Illuminates to indicate which step rescuer must take to treat a patient.
Shock Button Illuminates when the ZOLL AED Plus is charged and ready to deliver a
shock to the patient.
When pressed, causes the charged and ready ZOLL AED Plus to discharge its
energy into the patient.
When the ZOLL AED Plus is not charged the lighted button is extinguished.
Pressing this button initiates a voice prompt that indicates the number of
defibrillator shocks delivered since the unit was powered on.
Pictograms Icons that explain the series of steps needed for resuscitation and
defibrillation.
Status Indicator
Illuminated check mark indicates the unit passed its last self test and is ready
for use.
Illuminated “X” indicates unit has failed its self test and is not ready for use
LCD Display Displays elapsed time, shock count, user prompts, CPR compression depths
and ECG waveforms.
IrDA™ Port Provides a communications link between the defibrillator and a personal
computer or another IrDA™ equipped device.
PASS Cover Some ZOLL AED Plus models include a cover that may be used as a
(optional) shoulder support to aid patient airway management. The PASS can be
ordered separately for other ZOLL AED Plus models (see "Ordering
Accessories" on page 26).
Battery Holds ten (10) 123A lithium manganese dioxide batteries used to power the
Compartment unit.
Patient Connector Connector for attaching electrodes to the ZOLL AED Plus.
Speaker Provides audio prompts and metronome beeps that direct rescuers on what to
do during a resuscitation; also provides voice prompts to indicate if service is
required.
Microphone When voice recording option is installed, this microphone picks up and
(optional) records ambient sounds, including rescuer’s voice.
The ZOLL AED Plus unit contains an LCD display (some special models do not have an LCD) that
displays elapsed time, number of shocks delivered, text messages corresponding to the voice prompts,
and depth of CPR compressions. The LCD can also be configured to display the acquired ECG signals.
When the device is turned on, the unit automatically initiates the sequence of voice prompts and
graphic illuminations for a rescue event. The sequence continues until the device is turned off or the
electrodes are disconnected from the patient for an extended period of time. As soon as electrodes are
attached to the patient and impedance of the connection is verified, the unit stops cycling through the
above-mentioned voice and graphic illumination sequence and automatically begins analysis of the
ECG rhythm.
Following the results of this ECG analysis, voice prompts tell the rescuer whether a shockable or
non-shockable rhythm has been detected. If a shockable ECG rhythm is present, the graphics
illuminate and voice prompts guide the rescuer through the defibrillation sequence. When no shock is
advised, the ZOLL AED Plus issues the audio prompts, NO SHOCK ADVISED and START CPR, and
illuminates the CPR-related graphics. A period of 2 minutes (depending upon device configuration)
is then allowed for rescuer CPR. Following this “CPR period”, the ZOLL AED Plus automatically re-
initiates a new ECG rhythm analysis.
Voice Prompts
During clinical use of the ZOLL AED Plus, you may hear the following voice prompts.
Table 2: Clinical Voice Prompts
UNIT OK. ZOLL AED Plus has successfully passed its power up self tests.
UNIT FAILED. ZOLL AED Plus has failed its power up self tests and is not
usable for patient care.
CHANGE BATTERIES. ZOLL AED Plus self test has detected a low battery condition
that is insufficient for the device’s use for patient care. Replace
batteries immediately.
STAY CALM. Relax as much as possible and focus on the rescue effort.
CALL FOR HELP. Activate the EMS system or ask a bystander to do it for you.
OPEN AIRWAY. Place patient in the supine position and perform Head Tilt – Chin
Lift or Jaw – Thrust maneuver to open patient’s airway.
CHECK BREATHING. Look, listen or feel for the presence of breathing and/or airflow
from the patient’s lungs.
GIVE TWO BREATHS. If patient is not breathing, give two rescue breaths.
PLUG IN CABLE. Ensure that the electrode cable is properly connected to the
ZOLL AED Plus patient connector.
ATTACH DEFIB PADS TO Attach defibrillation pads to the patient’s bare chest.
PATIENT’S BARE CHEST.
CHECK ELECTRODE PADS. Previously attached electrodes are not making good contact with
the patient’s skin or the electrodes are defective.
ADULT PADS. ZOLL AED Plus has detected adult electrode pads connected to
it and adjusted defibrillation energy settings to adult levels.
PEDIATRIC PADS. ZOLL AED Plus has detected pediatric electrode pads connected
to it and adjusted defibrillation energy settings to pediatric levels.
DON’T TOUCH PATIENT, Do not touch patient, an ECG rhythm analysis is in progress or
ANALYZING. about to begin.
NO SHOCK ADVISED. ECG rhythm analysis has detected a rhythm that is not treatable
by defibrillation.
ANALYSIS HALTED. KEEP ECG rhythm analysis has been halted due to excessive ECG
PATIENT STILL. signal artifact. Stop any ongoing CPR and keep the patient as
motionless as possible.
DON’T TOUCH PATIENT. Warn all persons in attendance of the patient to stand clear and
PRESS FLASHING SHOCK stop touching the patient. Press the Shock Button to deliver
BUTTON. defibrillation therapy.
RELEASE SHOCK BUTTON. Shock Button was depressed before the defibrillator was ready to
defibrillate. Release the Shock Button and press again after the
ready tone sounds.
SHOCK DELIVERED. A defibrillation shock has just been delivered to the patient.
NO SHOCK DELIVERED. No shock was delivered to the patient because rescuer failed to
press Shock Button or an error condition was detected.
n SHOCKS DELIVERED. A total of n shocks have been delivered since the ZOLL AED
Plus was turned on.
CONTINUE CPR. Continue providing CPR. This prompt may also be issued if the
ZOLL AED Plus CPR monitoring function fails to detect chest
compressions at least ¾ of an inch deep.
PUSH HARDER. CPR compressions are consistently less than 1 ½ inches deep.
STOP CPR. Stop CPR, the ZOLL AED Plus is about to begin an ECG rhythm
analysis.
IF NEW BATTERIES, Press the Battery Reset Button located in the battery compartment
PRESS BUTTON. after replacing ALL batteries in the device with new batteries.
NON-RESCUE MODE. ZOLL AED Plus device has entered the diagnostics/data
communications mode.
COMMUNICATIONS IrDA Communications between the ZOLL AED Plus and a personal
ESTABLISHED. computer or modem have been established.
CPR Depth of
Shock Count Elapsed Time Compression Indicator
CHECK
1.5” RESPONSIVENESS 1.5” (3.8 cm)
(3.8 cm)
CHECK depth
2.0” (5.1 cm)
RESPONSIVENESS depth
2.0” (5.1 cm)
depth
Shock Count (Upper left corner of screen): Indicates the total number of defibrillation shocks
delivered by the ZOLL AED Plus since it was last powered on. Shock count is saved through brief
power off periods (< 5 seconds). When the ZOLL AED Plus is turned off for more than five (5)
seconds, the shock count is reset to 0.
Elapsed Time (Upper right corner of screen): Indicates the total time in minutes and seconds that has
elapsed since the ZOLL AED Plus was last powered on. Elapsed time continues to be counted through
brief power off periods (< 5 seconds). When the ZOLL AED Plus is turned off for more than five (5)
seconds, elapsed time is reset to 00:00. When elapsed time exceeds 99 minutes and 59 seconds, the
timer wraps around to 00:00 and continues counting.
CPR Depth of Compression Indicator (Right side of screen): A bar graph is displayed that shows
the depth of chest compressions measured during the delivery of CPR. Indicator lines are displayed in
the bar graph area at 1 ½ and 2 inches of compression depth to provide reference points for rescuers
performing CPR.
Visual User Prompts (Lower 1/3 of screen): Whenever the ZOLL AED Plus issues a voice prompt,
the text of the voice prompt is simultaneously displayed on the LCD display.
ECG Waveform (Center portion of screen): Although ZOLL AED Plus units do not display ECG
waveforms in their factory default configuration, the device can be specifically set up to continuously
display ECG signals as they are acquired.
WARNING! DO NOT use Passive Airway Support System (PASS) if suspected head or neck
injury. Place patient on a firm surface before performing CPR.
If the patient requires airway support and there is no evidence of head or neck trauma, roll the patient
on his/her side and then roll him/her back over so that the PASS is under the patient’s shoulders causing
the head to tilt backwards.
For PASS COVERS only: The shape of the PASS, when placed under the shoulders of the patient, can
be used to help maintain an open airway (see Figure 4).
The unit cover also functions as the Passive Airway Place the PASS under the patient to lift his/her shoulders.
Support System (PASS). Do not use the PASS if there is a suspected head or neck
injury.
The ZOLL AED Plus supports both adult and pediatric electrode pads. The device automatically
adjusts defibrillation energy to adult or pediatric levels depending on the type of electrodes connected
to it. Make sure that the electrode pads used are appropriate for the patient.
WARNING! DO NOT use adult electrode pads or CPR-D padz on patients under 8 years of age.
The ZOLL AED Plus uses electrode packs that are connected to the unit by a cable. The package
contains electrodes that you attach to the patient.
• Make sure to install a new package of electrodes and connect the electrode cable to the unit after
each use, to prepare for future emergencies.
• Check the electrode expiration date regularly to ensure that electrodes are fresh and ready to use
in an emergency situation.
• Replace electrodes, if expired
• After the ZOLL AED Plus is powered on and completes its self-test, the unit will issue an “Adult
Pads” voice prompt or a “Pediatric Pads” voice prompt to indicate the type of electrodes connected
to the unit. Verify that the connected electrode pads are appropriate for the patient being treated.
Connect alternate electrode pads if necessary.
If the electrodes are not attached properly, the unit issues the CHECK ELECTRODE PADS or the
ATTACH ELECTRODE PADS voice prompt during operation. If the electrode cable is not properly
attached to the unit, the unit issues the PLUG IN CABLE prompt. Make certain to connect the electrode
cable to the ZOLL AED Plus unit and attach the electrodes to the patient properly.
WARNING! Electrodes must be pre-attached to the device. Keep the electrode cable connected
to the ZOLL AED Plus unit at all times.
NOTE Electrodes contain no hazardous materials and may be disposed of in general trash unless contaminated
with pathogens. If contaminated, appropriate precautions should be used in their disposal.
WARNING! CPR-D padz are intended for use on adult patients only; do not use on patients
under 8 years of age.
NOTE If the patient is large or there is a need to place the electrode under a breast, you may need to tear away the
lower pad at the perforated line (see Figure 5) and extend the pad. Place the pad slightly to the patient’s left
and below the patient’s left breast.
NOTE If the patient has an implanted pacemaker or defibrillator in the upper right chest, angle the electrodes
slightly to avoid placing the electrodes over either device. Make certain that the CPR sensor maintains a
position over the lower half of the breastbone.
NOTE The Pedi Padz II (infant/child electrodes) can also be used with ZOLL pacemaker products for up to one
hour of pacing (see the M Series Operator’s Guide for information about pacing).
TM
II
Infant/Child Defibrillation Electrodes
0-8 Years <55lbs / 25kg
3
2
0197
ZOLL Medical Corporation 269 Mill Road
Chelmsford, MA 01824-4105 USA 978-421-9655
9305-0627-01 Rev. 1
NOTE If the ZOLL AED Plus prompts you to PUSH HARDER, your compressions are less than 1½ inches deep.
Increase your compression depth to improve CPR performance.
4. Deliver the appropriate number of rescue breaths then resume chest compressions. The metronome
will begin to beep again after your first few compressions have been delivered.
NOTE The unit allows for up to 3 minutes of audio recording prior to electrode attachment. When you turn the
unit off, the first indicator light (LED) on the graphical user interface illuminates and the second indicator
light (LED) flashes intermittently while the unit stores the data in memory.
Whenever the device detects a clinical event in rescue mode, the unit deletes previously stored data
(ECG, Audio and Event) from memory before data for the current rescue is recorded. Overwriting of
old ECG, audio and event data begins 10 seconds after the electrodes are properly connected to the
patient. However, if the unit is started in Non-rescue Mode, the recorded data of the last rescue is
retained and may be uploaded to a data storage or archiving system.
NOTE If the electrodes are not connected to the ZOLL AED Plus unit, the device will fail the self test and
display a red “X” in the status indicator window.
5. If the status indicator displays a red X, install new batteries. (See "Installing or Replacing
Batteries" on page 18.)
6. Close the top cover of the ZOLL AED Plus unit and initiate a self test by pressing the Power
Button. Verify that the unit issues the UNIT OK voice prompt. This prompt indicates that the new
batteries and electrodes are properly installed and that the unit is ready for service.
7. Verify that the ZOLL AED Plus unit issues the appropriate “Adult Pads” or “Pediatric Pads” voice
prompt.
8. Turn the ZOLL AED Plus unit off.
9. Wait 2 minutes. Verify that the green check symbol ( ) appears in the status indicator window
and that the unit does not emit a beeping tone.
10. Place the ZOLL AED Plus unit in service.
11. Check the ZOLL AED Plus unit periodically to ensure that the green check symbol ( ) appears
in the status indicator window.
NOTE If the status indicator displays a red X following completion of the above test, the ZOLL AED Plus is
not ready for use and may be defective. Remove the ZOLL AED Plus from service and consult the
Troubleshooting section on page 23 of this guide to help determine the problem.
CAUTION! DO NOT press the Battery Reset Button unless all batteries are new. Pressing the Battery Reset
Button when used batteries are installed may result in a false high reading of battery capacity. See
the "Installing or Replacing Batteries" on page 18 for more information.
SHOCKS: 00 00:12
Boot File Version
B:03.31 P:05.10 Application Software Version
Language File Version L:05.00 C:02.00 Firmware Version
H:10.00
Hardware Version
This self test verifies the following ZOLL AED Plus functions:
1. Battery Capacity: Verifies that the battery usage indicator shows at least 50% of full battery
capacity remaining.
2. Defibrillation Electrodes Connection: Verifies that the defibrillation electrodes are properly pre-
connected to the device.
3. ECG Circuitry: Verifies that the ECG signal acquisition and processing electronics are functional.
4. Defibrillator Charge and Discharge Circuitry: Verifies that the device’s defibrillator electronics
are functional and can charge and discharge at 2 joules.
5. Microprocessor Hardware/Software: Verifies proper function of the ZOLL AED Plus
microprocessor electronics and the integrity of its software.
6. CPR Circuitry and Sensor: Verifies that CPR monitoring and compression depth detection are
functional.
7. Audio Circuitry: Verifies that voice prompts are functional.
8. Display: Verifies that visual indicators are functional.
CAUTION! Use Duracell, Sanyo or Varta batteries only. Do not use Panasonic or Rayovac batteries.
Use of Panasonic or Rayovac batteries may result in significantly longer defibrillator
charging times than those required during emergency situations.
These batteries:
• Should be used well before their labeled expiration date.
• Should be checked periodically for the expiration date.
Batteries produced by all recommended manufacturers have a 10 year from date of manufacture shelf
life when not installed in a ZOLL AED Plus unit.
The following examples demonstrate how to read date codes on Duracell, Sanyo and Varta batteries.
Duracell:
The first digit indicates the year of manufacture (Example: 7=1997, 6=1996 etc.).
The second character is a letter from A-L that indicates the month of manufacture (Example:
A=January, B=February and so on to L for December.)
Sanyo:
The first letter indicates the year of manufacture (Example: A=1996, B=1997, etc.).
The second letter indicates the month of manufacture (Example: A=January, B=February, etc.).
Varta:
The BRAUNSCHWEIG - Code (always two letters) is printed on the side of the can in axial direction
to cell height.
The first letter indicates the month of manufacture (Example: B=January, R=February, A=March, etc.).
The second letter indicates the year of manufacture (Example: B=2001, R=2002, A=2003, etc.).
To install the batteries:
Battery
Compartment
Door
1) Depress latch
here
2. Remove all batteries at once and dispose of batteries properly. Place new batteries into the battery
bank, observing battery polarity markings and making sure that all batteries are securely seated
and properly oriented. After placing the first 5 to 9 batteries in the battery well, the
INSTALL BATTERIES audio prompt reminds you to install the remaining batteries in the battery
compartment.
CAUTION! You MUST replace all ten (10) batteries at once. Do not replace individual batteries. The
unit cannot detect whether all batteries or only a few batteries have been replaced. Do not
place used batteries into the unit. Using less than fully charged batteries can affect the
unit when performing a rescue. DO NOT press the Battery Reset Button if all the
batteries are not new. The unit then assumes that they are the same batteries that were just
removed.
NOTE If you do not press the Battery Reset Button in the battery well within 15 seconds after installing all
batteries, the device assumes that the batteries installed in the device were temporarily removed, and are
not fully charged.
NOTE Because Lithium Manganese Dioxide battery cells do not contain toxic materials, they may be disposed of
in general trash after discharge or when properly protected against shorting between terminals.
Low Battery with unit off. Audible beep from unit once Replace batteries.
every minute.
Low Battery or other self Status Indicator has red “X” Replace batteries. Check or
test failure with unit indicating failure to operate replace electrodes. If red “X”
powered off or during self (when off). remains, return to ZOLL
test. Technical Support for service.
Low Battery with unit CHANGE BATTERIES prompt Replace batteries as soon as
powered on. (unit powered on). possible.
Dead Battery. Status Indicator has red “X” Replace batteries. If red “X”
indicating failure to operate remains, return to ZOLL
(when off). Technical Support for service.
Maintenance Checklist
Use the following maintenance checklist when you periodically check your unit.
Table 4: Maintenance Checklist
Verify electrodes are connected to the unit and sealed in their package.
Replace if expired.
Are all cables free of cracks, cuts and exposed or broken wires?
Turn the unit on and off and verify the green check indicates ready for use.
Table 5: Troubleshooting
Self test failed. Initiate manual self test by pressing and holding the
ON/OFF Button for more than 5 seconds. Attempt
to repair the device by replacing the batteries or
electrodes. If unit fails test again, remove unit from
service and contact ZOLL Technical Service.
CHANGE BATTERIES prompt. Replace all batteries with new batteries at the same
time. Press the Battery Reset Button when
prompted.
Red “X” in status indicator window. Initiate manual self test by pressing and holding the
ON/OFF Button for more than 5 seconds.
Check to see if cable is attached properly to unit or
replace the electrodes.
Cycle power on the ZOLL AED Plus unit by
turning the unit off, then on again.
Replace all batteries at the same time with new
batteries. Press the Battery Reset Button when
prompted.
If unit still does not operate correctly, remove unit
from service and contact ZOLL Technical Service.
Beeping noise when unit is off. Remove unit from service and replace batteries.
Replace all batteries at the same time with new
batteries. Press the Battery Reset Button when
prompted.
If beeping continues, contact ZOLL Technical
Service.
ANALYSIS HALTED. KEEP PATIENT Excessive artifact detected during ECG analysis.
STILL prompt. Patient must be motionless during ECG analysis.
Do not touch the patient during analysis. Keep the
patient still. If the rescuer is using the device in an
emergency vehicle, bring the vehicle to a halt
before performing ECG analysis.
RELEASE SHOCK BUTTON prompt. Release Shock Button, then press and hold Shock
Button until discharge occurs. If voice prompt
continues, contact ZOLL Technical Service.
Have the following information available for the Technical Service representative:
• Unit serial number.
• Description of the problem.
• Purchase Order or credit card number to allow tracking of loan equipment.
• Purchase Order or credit card number for a unit with an expired warranty.
If you need to send the unit to ZOLL Medical Corporation, obtain a service order request number from
the Technical Service representative. Units are available on loan at an additional cost while your unit is
being repaired.
Remove all batteries from the unit and return the unit and batteries in its original container (or
equivalent packaging) with the service order request number on it to the following address:
ZOLL Medical Corporation
269 Mill Road
Chelmsford, MA 01824-4105
Attn: Technical Service Department
International Customers
Customers outside of the United States should remove all batteries from the unit and return the unit and
batteries in its original container (or equivalent packaging) to the nearest authorized ZOLL Medical
Corporation Service Center. To locate an authorized service center, contact the nearest ZOLL Sales
office or authorized distributor.
DEVICE
ENVIRONMENT
DEFIBRILLATOR
The ZOLL AED Plus is intended for use in the electromagnetic environment specified below.
The customer or the user of the ZOLL AED Plus should assure that it is used in such an
environment.
Medical Electrical Equipment needs special precautions regarding EMC and needs to be installed and
put into service according to EMC information provided in this document.
The ZOLL AED Plus is intended for use in the electromagnetic environment specified below.
The customer or the user of the ZOLL AED Plus should assure that it is used in such an
environment.
NOTE Ut is the a.c. mains voltage prior to application of the test level.
The ZOLL AED Plus is intended for use in the electromagnetic environment specified below.
The customer or the user of the ZOLL AED Plus should assure that it is used in such an
environment.
The ZOLL AED Plus is intended for use in an environment in which radiated RF disturbances are controlled. The
customer or the user of the ZOLL AED Plus can help prevent electromagnetic interference by maintaining a
minimum distance between portable and mobile RF communications equipment (transmitters) and the ZOLL
AED Plus as recommended below, according to the maximum output power of the communications equipment.
Rated maximum 150 kHz to 80 150 kHz to 80 80 MHz to 800 800MHz to 2.5
output power of MHz outside MHz in ISM MHz GHz
transmitter ISM bands bands
W d =[ 3.5
3
] √P d = [ 12
10 √P
] d = [ 12
10 √P
] d =[ 23 ]
10 √P
For transmitters rated at a maximum output power not listed above, the recommended separation
distance d in metres (m) can be determined using the equation applicable to the frequency of the
transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to
the transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2 The ISM (industrial, scientific and medical) bands between 150 kHz and 80 MHz are 6.765
MHz to 6.795 MHz; 13.553 MHz to 13.567 MHz; 26.957 MHz to 27.283 MHz; and 40.66 MHz to
40.70 MHz.
NOTE 3 An additional factor of 10/3 is used in calculating the recommended separation distance for
transmitters in the ISM frequency bands between 150 kHz and 80 MHz and in the frequency range
80 MHz to 2.5 GHz to decrease the likelihood that mobile/portable communications equipment
could cause interference if it is inadvertently brought into patient areas.
NOTE 4 These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.
First Phase 32 A 26 A 21 A 17 A
Maximum Initial
Current
Interphase duration 150 µsec 150 µsec 150 µsec 150 µsec
between first and second
phases
The efficacy of ZOLL's Rectilinear Biphasic Waveform has been clinically verified during a
Ventricular Fibrillation (VF) and Ventricular Tachycardia (VT) defibrillation study. This study (which
was conducted using ZOLL M Series defibrillators) and the findings are described below. Since the
ZOLL AED Plus's rectilinear biphasic waveform employs the same first and second phase timing,
similar first and second phase currents/voltages and essentially the same mechanisms for controlling
defibrillation waveshape, the ZOLL M Series and ZOLL AED Plus defibrillation waveforms are
considered substantially equivalent.
Monophasic Biphasic
p-value 0.0517
Successful defibrillation with rectilinear biphasic shocks was achieved with 58% less delivered
current than with monophasic shocks (14 ±1 vs. 33 ±7 A, p=0.0001).
Monophasic Biphasic
p-value 0.02
A single patient required a second biphasic shock at 150J to achieve 100% efficacy versus six patients
for whom monophasic shocks of up to 360J were required for 100% total defibrillation efficacy.
Conclusion: The data demonstrate the equivalent efficacy of low energy rectilinear biphasic shocks
compared to standard high energy monophasic shocks for transthoracic defibrillation for all patients
at the 95% confidence level. The data also demonstrate the superior efficacy of low energy rectilinear
biphasic shocks compared to standard high energy monophasic shocks in patients with high
transthoracic impedance at the 90% confidence level. There were no unsafe outcomes or adverse
events due to the use of the rectilinear biphasic waveform.
* Kerber, R., et. al., AHA Scientific Statement, Circulation, 1997; 95: 1677-1682:
“... the task force suggests that to demonstrate superiority of an alternative waveform over standard
waveforms, the upper boundary of the 90% confidence interval of the difference between standard and
alternative waveforms must be < 0% (i.e., alternative is greater than standard).”
If an AED Plus unit is equipped for treating adult and pediatric patients, when you
connect the cable for adult electrodes, the AED Plus unit switches automatically to adult mode for cardiac
rhythm detection and automatic shock energy selection (120 Joules, 150 Joules, 200 Joules) and issues the
voice and display prompt , ADULT PADS. When you connect the cable for ZOLL pedi•padz II pediatric
electrodes, the AED PLUS unit switches automatically to pediatric mode for cardiac rhythm detection and
automatic shock energy selection (50 Joules, 70 Joules, 85 Joules) and issues the voice and display
prompt, PEDIATRIC PADS.
NOTE You can also use pedi•padz II (infant/child electrodes) with ZOLL pacemaker products for up to
one hour of pacing (see the M Series Operator’s Guide for information about pacing).
Unconsciousness; and
Absence of normal breathing; and
Absence of a pulse or signs of circulation.
When the victim is less than 8 years of age and weighs less than 55 lbs (25 kg), the ZOLL AED Plus should
be used with the ZOLL AED Plus Pediatric Electrodes. Therapy should not be delayed to determine the
patient’s exact age or weight.
9652-0839-01 Rev. B
Warning! Set-up and Check-out Procedure:
Use the AED Plus unit only as described in this manual. Improper use of the 1. Insert 10 new batteries into AED Plus unit.
device can cause death or injury. 2. Connect electrode cable to AED Plus unit and pack sealed electrodes
DO NOT use or place the AED Plus unit in service until you have read the AED inside unit cover. Close cover.
Plus Operator’s and Administrator’s Guides. 3. Turn unit on and wait for “Unit OK” audio message. Verify that unit
DO NOT use or place the AED Plus unit in service if the unit’s status indicator
issues appropriate “Adult Pads” or “Pediatric Pads” audio message.
window (located on the left side of the handle) displays a red “X”. 4. Turn unit off.
5. Wait 2 minutes. Verify that green check symbol ( ) appears in status
DO NOT use or place the AED Plus unit in service if the unit emits a beeping tone.
indicator window (located on left side of handle) and that unit does not
Connect the electrode cable to the AED Plus unit after installing batteries. emit a beeping tone.
Keep the electrode cable connected to the AED Plus unit at all times. 6. Place AED Plus unit in service.
This device should only be used by properly trained individuals. 7. Check AED Plus unit periodically to ensure that green check symbol ( )
Only use electrodes labeled “Infant/Child” on children less than 8 years old or appears in status indicator window.
weighing less than 55 lbs (25 kg). Use CPR-D padz if patient is older than 8 years
or weighs more than 55 lbs (25 kg).
Battery Replacement
Always stand clear of patient when delivering treatment. Defibrillation energy Replace batteries before expiration date or if unit prompts. Use only type
delivered to the patient may be conducted through the patient’s body and cause
a lethal shock to those touching the patient. 123A lithium manganese dioxide batteries from recommended manufacturers.
DO NOT touch the electrode surfaces, the patient, or any conductive material Remove all batteries from battery compartment and discard before
touching the patient during ECG analysis or defibrillation. installing any new batteries.
Insert 10 new batteries into battery well. Do not use old batteries.
Move patient away from electrically conductive surfaces prior to use of equipment.
Press button in battery well only after installation of new batteries.
DO NOT use the unit near or within puddles of water.
Keep the patient as motionless as possible during ECG analysis.
DO NOT use the unit near flammable agents, such as gasoline, oxygen-rich
Cleaning
atmospheres, or flammable anesthetics. Clean and disinfect unit with soft, damp cloth using 90% isopropyl
Avoid radio frequency interference from high-power sources that might cause the alcohol or soap and water, or chlorine bleach (30 ml/liter water).
defibrillator to interpret cardiac rhythms incorrectly by turning off cell phones and Do not immerse any part of the unit in water.
2-way radios. Do not use ketones (MEK, acetone, etc.).
Avoid using abrasives (e.g., paper towels) on the LCD display, if so
Disconnect non-defibrillation protected electronic devices or equipment from
patient before defibrillation. equipped.
Do not sterilize the unit.
Dry victim’s chest, if wet, before attaching electrodes.
Apply freshly opened and undamaged electrodes, within the electrode expiration
date, to clean and dry skin to minimize burning.
DO NOT place electrodes directly over the patient’s implanted pacemaker.
Pacemaker stimuli may degrade the accuracy of ECG rhythm analyses or the
TROUBLESHOOTING Automated External Defibrillator
Operator’s Guide
pacemaker may be damaged by defibrillator discharges.
Check labeling inside the ZOLL AED Plus cover before using the cover as a Passive
Airway Support System (PASS) device to ensure it is intended for this use.
Problem Recommended Action
DO NOT use the Passive Airway Support System (PASS) if there is a suspected
head or neck injury. Place the patient on a firm surface before performing CPR.
Self-test failed. Manually test by pressing and AHA/ERC 2005
DO NOT recharge, disassemble, or dispose of batteries in fire. Batteries may
explode if mistreated. holding the ON/OFF button for
more than 5 seconds. If unit fails
DO NOT use or stack the AED PLUS unit with other equipment. If the unit is used test again, remove from service.
or stacked with other equipment, verify proper operation prior to use.
Caution!
“Change batteries” Replace all batteries at the For technical support or repair:
prompt. same time.
DO NOT disassemble the unit. A shock hazard exists. Refer all servicing to Technical Service Department
qualified personnel. Red “X” in Status Perform manual test.
Indicator window Check to see if cable is ZOLL Medical Corporation
Use only commercially available type 123A lithium manganese dioxide batteries. OR attached properly to unit.
Discard batteries properly after removal from unit. Use only batteries from 269 Mill Road
recommended manufacturers. See the AED Plus Administrator’s Guide (P/N beeping noise Replace batteries. Chelmsford, MA 01824-4105
9650-0301-01) for a list of recommended battery manufacturers. when unit is OFF. If unit still does not operate 978-421-9655 • 800-348-9011
correctly, remove from service. Fax: 978-421-0010
If the device is stored outside the recommended environmental conditions, the
electrode pads and/or batteries may be damaged or their useful life reduced.
Red “X” in Status Indicator Power cycle the unit. If Red “X”
The CPR-D Padz Electrode can be connected to other ZOLL defibrillators with window when unit is ON. is still present in Status Indicator International Service: Contact your local distributor.
Multifunction Cables. Defibrillation can be administered when connected to window, remove unit from
other ZOLL defibrillators.The CPR function does not operate with any device
service.
other than the AED Plus defibrillator.
Important!
This symbol indicates that an AED Plus unit is equipped for treating adult and
pediatric patients. An AED Plus unit without this symbol is not equipped to treat
pediatric patients and will NOT work with pedi•padz II ™ pediatric electrodes. To Federal (U.S.A.) law restricts this device for sale to or on the order of a physician. © ZOLL Medical Corp. Inc. 2006
upgrade an AED Plus unit for use with ZOLL pedi•padz II pediatric electrodes,
Conformité Européene - Complies with the Medical Device Directive 93/42/EEC REF 9650-0300-01 Rev. H
contact ZOLL Medical Corporation or an authorized ZOLL distributor for
information on the ZOLL AED Plus Pediatric Upgrade Kit.
Indicator light flashes. If unresponsive, Open Airway.
Check responsiveness.