Flow Measurement
Flow Measurement
Flow Measurement
COM
VOLUME
4
Click here to visit the OMEGA.COM website
TABLE OF CONTENTS
Turbulent
Laminar
Velocity
Velocity
or
• The Flow Pioneers Flow
Profile
Flow
Profile
3 Flow
Figure 3-7:
1.00
K = 1 Asymptote
0.90
For Flat Profile
Re
Figure 4-6:
Support
Mass Flowtube
(Typical)
Enclosure
Mass Tube Enclosure
Mass Flowmeters
Pipe/Flowtube
Figure 5-5:
04 Volume 4 TRANSACTIONS
REFERENCE SECTIONS
Editorial 06 106 Information Resources
About OMEGA 07 110 Glossary
Level %
100
• Level Sensor Selection Vertical
Sphere
Figure 6-3:
Bimetallic
Pneumatic
Range
Nozzle & Flapper Relay
Liquid Filled
Diaphragm
Side
Capsule
A) B)
Figure 7-3:
Level Kv
#1 #2 C= KA
D
D
• Theory of Operation - --
-
- --
Kl C=Capacitance
K=Dieletric Constant
A=Area of Plates
- -- -
-- + ++ D=Dist. Between Plates
8 RF/Capacitance Level Instrumentation • Probe Designs
-
- --
-
- --
A
+ ++
++
++
++
+
Voltmeter Electron
Flow 87
-- ++ A
+ ++ RF
- +
+ ++ +
• Installation Considerations ++
++
+
Ammeter
A) B)
Figure 8-2:
Reflection
A) B)
Figure 9-6:
Figure 10-4:
TRANSACTIONS Volume 4 05
1
80
20
Orifice Plate while decreasing its static energy. It is
for this reason that a flow restriction
70
30
Recovery—Percent of Differential
08 Volume 4 TRANSACTIONS
1 A Flow Measurement Orientation
displacement and turbine meters, where C is the constant for units single, dimensionless ratio to describe
about one; and, through flow venturis, conversion. the velocity profile of flowing fluids:
less than 0.5 heads. Therefore, if an ori- Over the past several years, the
fice plate (Figure 1-2) with a beta ratio performance of magnetic flowmeters Re = DVρ/µ
TRANSACTIONS Volume 4 09
A Flow Measurement Orientation 1
are used at very low flowrates (capil- flowmeter is similar. The French civil vortex flowmeter, which determines
lary flowmeters) or when the viscos- engineer Gaspard Coriolis discovered flow velocity by counting the num-
ity of the process fluid is high. in 1843 that the wind, the ocean cur- ber of vortices passing a sensor. Von
In the case of some flowmeter rents, and even airborne artillery Karman published his findings in
technologies, more than a century shells will all drift sideways because 1954, and because by that time the
elapsed between the discovery of a of the earth’s rotation. In the northern sensors and electronics required to
hemisphere, the deflection is to the count vortices were already in exis-
Conventional
right of the motion; in the southern tence, the first edition of the
Magnetic Flowmeters hemisphere, it is to the left. Similarly, Instrument Engineers’ Handbook in
Performance of Pulsed
a body traveling toward either pole 1968 was able to report the availabil-
DC Magnetic Flowmeters will veer eastward, because it retains ity of the first swirlmeter.
4.0 the greater eastward rotational speed The computer has opened new
3.0 of the lower altitudes as it passes frontiers in all fields of engineering,
% Rate Accuracy
2.0 over the more slowly rotating earth and flow measurement is no excep-
1.0 surface near the poles. Again, it was tion. It was only as long ago as 1954
0.5
0 the slow evolution of sensors and that another Hungarian-American
-0.5
-1.0 electronics that delayed creation of mathematician, John Von Neumann,
-2.0 the first commercial Coriolis mass built Uniac—and even more recently
-3.0 flowmeter until the 1970’s. that yet another Hungarian-
-4.0 It was the Hungarian-American American, Andy Grove of Intel,
10 50 100 aeronautical engineer Theodore developed the integrated circuit. Yet
% Full Scale von Karman who, as a child growing these events are already changing
Figure 1-4: Magmeter Accuracy
up in Transylvania (now Romania), the field of flowmetering. Intelligent
noticed that stationary rocks caused differential pressure cells, for exam-
scientific principle and its use in vortices in flowing water, and that ple, can automatically switch their
building a flowmeter. This is the case the distances between these travel- range between two calibrated spans
with both the Doppler ultrasonic and ing vortices are constant, no matter (one for 1-10%, the other for 10-100%
the Coriolis meter. how fast or slow the water runs. of D/P), extending orifice accuracy
In 1842, the Austrian physicist Later in life, he also observed that, to within 1% over a 10:1 flow range.
Christian Doppler discovered that, if a when a flag flutters in the wind, the Furthermore, it is possible to include
sound source is approaching a receiver wavelength of the flutter is indepen- in this accuracy statement not only
(such as a train moving toward a sta- dent of wind velocity and depends hysteresis, rangeability, and linearity
tionary listener), the frequency of the
sound will appear higher. If the source
and the recipient are moving away
from each other, the pitch will drop
(the wavelength of the sound will
appear to decrease). Yet it was more
than a century later that the first ultra-
sonic Doppler flowmeter came on the
market. It projected a 0.5-MHz beam
into a flowing stream containing reflec-
tors such as bubbles or particles. The
shift in the reflected frequency was a Flow measurement options run the gamut from simple, economical paddle wheels (shown) to
function of the average traveling veloc- sophisticated high-accuracy devices.
ity of the reflectors. This speed, in turn,
could be used to calculate a flowrate. solely on the diameter of the flag effects, but also drift, temperature,
The history of the Coriolis pole. This is the theory behind the humidity, vibration, over-range, and
10 Volume 4 TRANSACTIONS
1 A Flow Measurement Orientation
power supply variation effects. Those “inexpensive” purchases can be data be filled in for each application:
With the development of super- the most costly installations. • Fluid and flow characteristics: In
chips, the design of the universal The basis of good flowmeter this section of the table, the name
flowmeter also has become feasible. selection is a clear understanding of of the fluid is given and its pressure,
It is now possible to replace dye- the requirements of the particular temperature, allowable pressure
tagging or chemical-tracing meters application. Therefore, time should drop, density (or specific gravity),
(which measured flow velocity by be invested in fully evaluating the conductivity, viscosity (Newtonian
dividing the distance between two nature of the process fluid and of the or not?) and vapor pressure at
points by the transit time of the overall installation. The development maximum operating temperature
trace), with traceless cross-correla- of specifications that state the appli- are listed, together with an indica-
tion flowmeters (Figure 1-6). This is
an elegant flowmeter because it
Coefficient of Discharge
requires no physical change in the
process—not even penetration of Concentric
Square-Edged
the pipe. The measurement is based Orifice Magnetic
on memorizing the noise pattern in Flowmeter
any externally detectable process Eccentric
Orifice
variable, and, as the fluid travels
from point A to point B, noting its
Integral
transit time. Target Meter
Orifice
(Best Case) Flow
Venturi Tube
Nozzle
Flow Sensor Selection Target Meter
The purpose of this section is to (Worst Case) Quadrant-Edged
Pipeline
Orifice Reynolds
provide information to assist the Number
reader in making an informed selec-
10 102 103 104 105 106
tion of flowmeter for a particular
application. Selection and orienta-
Figure 1-5: Effect of Reynolds Numbers on Various Flowmeters
tion tables are used to quickly focus
on the most likely candidates for cation requirements should be a sys- tion of how these properties
measurement. Tables 1-I and 1-II tematic, step-by-step process. might vary or interact. In addition,
have been prepared to make avail- The first step in the flow sensor all safety or toxicity information
able a large amount of information selection process is to determine if should be provided, together with
for this selection process. the flowrate information should be detailed data on the fluid’s compo-
At this point, one should consider continuous or totalized, and whether sition, presence of bubbles, solids
such intangible factors as familiarity of this information is needed locally or (abrasive or soft, size of particles,
plant personnel, their experience with remotely. If remotely, should the fibers), tendency to coat, and light
calibration and maintenance, spare transmission be analog, digital, or transmission qualities (opaque,
parts availability, mean time between shared? And, if shared, what is the translucent or transparent?).
failure history, etc., at the particular required (minimum) data-update fre- • Expected minimum and maximum
plant site. It is also recommended that quency? Once these questions are pressure and temperature values
the cost of the installation be comput- answered, an evaluation of the prop- should be given in addition to the
ed only after taking these steps. One erties and flow characteristics of the normal operating values. Whether
of the most common flow measure- process fluid, and of the piping that flow can reverse, whether it does
ment mistakes is the reversal of this will accommodate the flowmeter, not always fill the pipe, whether
sequence: instead of selecting a sensor should take place (Table 1-I). In order slug flow can develop (air-solids-liq-
which will perform properly, an to approach this task in a systematic uid), whether aeration or pulsation
attempt is made to justify the use of a manner, forms have been developed, is likely, whether sudden tempera-
device because it is less expensive. requiring that the following types of ture changes can occur, or whether
TRANSACTIONS Volume 4 11
A Flow Measurement Orientation 1
special precautions are needed dur- with sanitary or clean-in-place sold or purchased on the basis of a
ing cleaning and maintenance, these (CIP) regulations. meter reading, absolute accuracy is
facts, too, should be stated. The next step is to determine the critical. In other applications,
• Concerning the piping and the area required meter range by identifying repeatability may be more important
where the flowmeter is to be locat- minimum and maximum flows (mass than absolute accuracy. Therefore, it
ed, the following information or volumetric) that will be measured. is advisable to establish separately
the accuracy and repeatability
nB e requirements of each application and
s i tio rt Pip to state both in the specifications.
Po p o
ns
Tra When a flowmeter’s accuracy is
lay
e De stated in % CS or % FS units, its
A Tim absolute error will rise as the mea-
ion
sit sured flow rate drops. If meter error is
Po
stated in % AR, the error in absolute
n(t)
w terms stays the same at high or low
Flo
m(t) flows. Because full scale (FS) is always
A a larger quantity than the calibrated
span (CS), a sensor with a % FS perfor-
Upstream
mance will always have a larger error
m(t) Transducer Signal than one with the same % CS specifi-
cation. Therefore, in order to compare
all bids fairly, it is advisable to convert
Time. t
all quoted error statements into the
Transit
same % AR units.
Time
It is also recommended that the
user compare installations on the
Downstream
basis of the total error of the loop. For
n(t) Transducer Signal
B example, the inaccuracy of an orifice
plate is stated in % AR, while the error
of the associated d/p cell is in % CS
Time. t or % FS. Similarly, the inaccuracy of a
Figure 1-6: The Ultrasonic Transit-Time Flowmeter Coriolis meter is the sum of two
errors, one given in % AR, the other as
should be specified: For the piping, After that, the required flow mea- a % FS value. Total inaccuracy is calcu-
its direction (avoid downward flow surement accuracy is determined. lated by taking the root of the sum of
in liquid applications), size, material, Typically accuracy is specified in per- the squares of the component inaccu-
schedule, flange-pressure rating, centage of actual reading (AR), in racies at the desired flow rates.
accessibility, up or downstream percentage of calibrated span (CS), or In well-prepared flowmeter specifi-
turns, valves, regulators, and avail- in percentage of full scale (FS) units. cations, all accuracy statements are
able straight-pipe run lengths. The accuracy requirements should be converted into uniform % AR units and
• In connection with the area, the separately stated at minimum, nor- these % AR requirements are specified
specifying engineer must know if mal, and maximum flowrates. Unless separately for minimum, normal, and
vibration or magnetic fields are pre- you know these requirements, your maximum flows. All flowmeter specifi-
sent or possible, if electric or pneu- meter’s performance may not be cations and bids should clearly state
matic power is available, if the area acceptable over its full range. both the accuracy and the repeatabili-
is classified for explosion hazards, ty of the meter at minimum, normal,
or if there are other special Accuracy vs. Repeatability and maximum flows.
requirements such as compliance In applications where products are Table 1 provides data on the range
12 Volume 4 TRANSACTIONS
1 A Flow Measurement Orientation
of Reynolds numbers (Re or RD) with- when flow and density are at their no moving parts, select the one
in which the various flowmeter maximum and viscosity at its mini- without moving parts. Moving parts
designs can operate. In selecting the mum. Conversely, the minimum RD is are a potential source of problems,
right flowmeter, one of the first steps obtained by using minimum flow and not only for the obvious reasons of
is to determine both the minimum density and maximum viscosity. wear, lubrication, and sensitivity to
and the maximum Reynolds numbers If acceptable metering performance coating, but also because moving
for the application. Maximum RD is can be obtained from two different parts require clearance spaces that
obtained by making the calculation flowmeter categories and one has sometimes introduce “slippage” into
HIGH TEMPERATURE
SLURRIES
NON-NEWTONIANS
VISCOUS
SEMI-FILLED
PIPES
PULSATING
FLOW
VERY CORROSIVE
PRESS
OPEN CHANNEL
REVERSE FLOW
CRYOGENIC
CLEAN
CORROSIVE
FIBROUS
ABRASIVE
TYPICAL
TYPICAL
STEAM
CLEAN
DIRTY
DIRTY
HIGH
LOW
LOW
FLOWMETER PIPE SIZE, in. (mm)
(Including transmitter)
or viscosity
°F (°C)
psig (kPa)
SQUARE ROOT SCALE: MAXIMUM SINGLE RANGE 4:1 (Typical)**
to -30-250°F (-30-120°C)
Orifice
Process temperature
Transmitter limited
to 1000°F (540°C):
√
√
X
√
√
√
X
?
?
X
X
X
X
SD
?
√
√
X
?
X
±1-4% URV
psig
Square-Edged
>1.5 (40)
RD > 10,000
(41,000 kPa)
Honed Meter Run
0.5-1.5 (12-40)
√
√
X
√
√
√
?
?
?
X
X
X
X
SD
?
√
√
X
?
X
±1% URV
R > 10,000
To 4,000
D
X
±2-5% URV
?
√
X
√
√
√
X
?
?
X
?
X
X
SD
?
?
X
X
?
Integrated
<0.5 (12)
RD > 10,000
Segmental Wedge
<12 (300)
√
√
√
√
√
√
?
?
√
?
X
?
?
SD
?
√
√
X
?
X
±0.5% URV
RD > 500
Eccentric
>2 (50)
?
?
√
√
√
?
X
?
?
?
X
?
X
SD
?
√
√
X
?
X
±2-4% URV
RD > 10,000
Segmental
>4 (100)
?
?
√
√
√
?
X
?
?
?
X
?
X
SD
?
√
√
X
?
X
±2-4% URV
RD > 10,000
V-Cone
0.5-72 (12-1800)
√
√
?
√
√
√
?
?
√
?
X
?
?
X
?
?
?
X
?
X
±0.5-1% of rate
RD : 8,000-5,000,000
700 (370)
≤600 (4,100)
Target***
<0.5(12)
?
√
√
√
√
√
?
?
√
√
X
X
X
?
X
?
?
X
?
X
±0.5-5% URV
RD > 100
to -30-250°F (-30-120°C)
Process temperature
Transmitter limited
X
±0.5-2% URV
to 1000°F (540°C):
Venturi
>2 (50)
√
√
?
√
√
√
?
?
√
?
X
√
?
X
?
?
?
X
?
RD > 75,000Ł
psig
(41,000 kPa)
Flow Nozzle
>2 (50)
?
√
?
√
√
√
X
?
?
?
X
X
X
X
?
?
?
X
?
X
±1-2% URV
RD > 50,000Ł
To 4,000
X
±1.25% URV
√
√
X
√
√
√
X
√
?
X
X
X
X
X
?
?
?
X
?
Low Loss Venturi
>3 (75)
RD > 12,800Ł
Pitot
>3 (75)
X
√
X
√
√
√
X
?
?
X
X
X
X
X
X
?
?
X
X
X
±3-5% URV
RD > 100,000Ł
Averaging Pitot
>1 (25)
√
√
SD
√
√
√
X
?
?
SD
X
X
X
SD
X
?
?
X
X
X
±1-2% URV
RD > 40,000Ł
Elbow
>2 (50)
X
√
?
√
√
√
X
?
?
?
X
X
X
√
X
?
?
X
?
X
±5-10% URV
RD > 10,000Ł
Laminar
0.25-16.6 (6-400)
?
√
X
√
√
√
√
?
?
X
X
X
X
X
√
X
X
X
X
X
±1% of rate
RD < 500
150 (66)
≤30 (225)
LINEAR
SCALE TYPICAL RANGE
10:1 (Or better)
Magnetic*
0.1-72 (2.5-1800)
X
X
X
X
X
√
?
√
√
√
√
√
√
√
√
?
X
?
?
?
±0.5% of rate
RD > 4,500
360 (180)
≤ 1,500 (10,800)
Positive Displacement
Gas
<12 (300)
X
√
X
?
√
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
±1% of rate
-
250 (120)
≤ 1,400 (10,000)
Liquid
<12 (300)
X
X
X
X
X
√
√
?
X
?
X
X
X
X
X
X
?
X
X
X
±0.5% of rate
No RD limit ≤ 8,000 cS
600 (315)
≤ 1,400 (10,000)
Turbine
Gas
0.25-24 (6-600)
SD
√
X
√
√
X
X
X
X
X
X
X
X
SD
SD
?
?
X
X
?
±0.5% of rate
-
-450-500 (268-260)
≤ 3,000 (21,000)
Liquid
0.25-24 (6-600)
X
X
X
X
X
√
X
?
X
?
X
X
SD
SD
SD
?
?
X
X
?
±0.5% of rate
Rp > 5,000, ≤15 cS
-450-500 (268-260)
≤ 3,000 (21,000)
Ultrasonic
Time of Flight
>0.5 (12)
X
SD
SD
SD
SD
√
?
?
X
√
√
?
?
√
√
X
?
X
X
?
±1% of rate to ±5% URV
RD > 10,000
-300-500 (-180-260)
Pipe rating
Doppler
>0.5 (12)
X
X
X
X
X
X
?
?
√
√
√
√
√
√
√
X
X
X
?
X
±1% of rate to ±5% URV
RD > 4,000
-300-500 (-180-260)
Pipe rating
Variable-Area (Rotameter)
≤3 (75)
?
√
X
X
√
√
X
?
X
?
?
X
X
X
?
?
?
X
X
X
±1% of rate to ±10% URV
No RD limit, < 100 cS
Glass: 400 (200)
Glass:
350 (2,400)
Metal: 1,000 (540)
Metal: 720 (5,000)
Vortex Shedding
1.5-16 (40-400)
√
√
?
√
√
√
X
?
?
?
X
X
X
X
X
?
?
X
X
X
±0.75-1.5% of rate
RD > 10,000, < 30 cP
400
(200)
≤ 1,500 (10,500)
X
±0.5% of rate
Vortex Precession (Swirl)
<16 (400)
√
√
?
√
√
√
X
?
X
?
X
X
X
X
X
?
X
X
X
RD > 10,000, < 5 cP
536
(280)
Pipe rating
Fluidic Oscillation (Coanda)
>1.5 (40)
X
X
X
X
X
√
X
X
?
?
X
X
X
X
?
?
?
X
X
X
±2% of rate
RD > 2,000, < 80 cS
350
(175)
≤ 720 (5,000)
Mass
Coriolis
0.25-6 (6-150)
?
?
?
√
√
√
√
√
√
?
?
?
√
?
?
?
?
X
√
X
±0.15-10% of rate
No RD limit
-400-800 (-224-427)
≤ 5,700 (39,900)
Thermal Probe
<72 (1800)
X
√
?
√
√
√
?
√
√
?
?
?
?
X
?
?
X
X
?
X
±1-2% URV
No RD limit
1,500 (816)
Pipe rating
Solids Flowmeter
<24 (600)
X
X
X
X
X
X
SD
X
?
X
X
SD
SD
X
SD
SD
X
√
X
X
±0.5% of rate to ±4% URV
-
750 (400)
≤ 580 (4,000)
Correlation
X
No data available
Capacitance
<8 (200)
X
X
X
X
X
X
√
√
√
√
√
√
√
X
?
?
X
?
?
No data available
300 (149)
≤ 580 (4,000)
Ultrasonic
>0.5 (12)
X
X
X
X
X
X
?
√
√
√
√
√
√
X
?
X
X
X
?
X
±6% of ??
No data available
-300-250 (-180-120)
Pipe rating
cP = centi Poise
? = Normally applicable (worth consideration)
URV = Upper Range Value
‡ According to other sources, the minimum
* Liquid must be electrically conductive
cS = centi Stokes
√ = Designed for this application (generally suitable)
X = Not applicable
Reynolds number should be much higher
** Range 10:1 for laminar, and 15:1 for target
SD = Some designs
*** Newer designs linearize the signal
TRANSACTIONS Volume 4 13
A Flow Measurement Orientation 1
the flow being measured. Even also change the internal dimensions of generally advisable to use the
with well maintained and calibrated the meter and require compensation. flowmeter. Because point sensors do
meters, this unmeasured flow varies Furthermore, if one can obtain the not look at the full flow, they read
with changes in fluid viscosity and same performance from both a full accurately only if they are inserted to
temperature. Changes in temperature flowmeter and a point sensor, it is a depth where the flow velocity is
(UPSTREAM DIAM./DOWNSTREAM DIAM.)
0.1
1.0
10
102
103
104
105
106
Units
VOLUME DISPLACEMENT-FLOW SENSOR
DIFFERENTIAL PRESSURE-FLOW SENSOR
cc/min
0.05
0.3
2.8
28.3
Sm
/hr or Am
/hr
3 3
Gas
PRESSURE LOSS THRU SENSOR
10
-6
-5
10 10
-4
10-3
10-2
0.1
1.0
10
102
103
104
Flow
DIRECT MASS-FLOW SENSOR
Units
TRANSMITTER AVAILABLE
-2
VELOCITY-FLOW SENSOR
10
-6
10
-5
10
-4
10
-3
10
0.1
1.0
10
102
103
104
105
cc/min
LINEAR OUTPUT
Liquid
10-6
10-5
10-4
10-3
10-2
0.1
1.0
10
102
103
104
m3/hr
Flow
Units
-4
-3
-2
10-6
10-5
10
10
10
0.1
1.0
10
102
103
104
105
106
gpm
TYPE OF DESIGN
gpm—m3/hr
➂
Orifice (plate or integral cell)
√
H
√
SR
3:1
➁ H
20/5
SCFM—Sm3/hr
M
√
SR
3:1
➁
3
Segmental Wedge
√
A
20/5
gpm—m /hr
➁
gpm—m
3/hr
V-Cone Flowmeter
√
√
SR
3:1 to 15:1
M
2/5
ACFM—Sm
3/hr
gpm—m3/hr
➃
Target Meters
√
A
√
SR
15:1
M
20/5
SCFM—Sm /hr
3
Venturi Tubes
√
H
√
SR
3:1
➁ M
15/5
3/hr
gpm—m
➂
SCFM—Sm3/hr
Flow Nozzles
√
H
√
SR
3:1
➁ A
20/5
gpm—m3/hr
➃
Pitot Tubes
√
M
√
SR
3:1
➁ M
30/5
3
/hr
SCFM—Sm
3/hr
➂
M
√
SR
3:1
➁
gpm—m
Elbow Taps
√
N
25/10
SCFM—Sm
3/hr
3
gpm—m /hr
Laminar Flowmeters
√
√
√
10:1
➁ H
15/5
SCFM—Sm3/hr
√
N
√
√
30:1
➆
3
Magnetic Flowmeters
N
5/3
gpm—m /hr
Positive Displacement
√
SD
√
10:1 to
M
N
3
SCFM—Sm /hr
Gas Meters
200:1
➆
Positive Displacement
√
M
SD
√
10:1
A
N
gpm—m
3
/hr
Liquid Meters
gpm—m3/hr
Turbine Flowmeters
√
√
H
√
√
10:1
➇ A
15/5
SCFM—Sm3/hr
Ultrasonic Flowmeters
Time of Flight
√
√
N
√
√
20:1
N
20/5
gpm—m3/hr
➂
Doppler
√
√
N
√
√
10:1
N
20/5
SCFM—Sm3/hr
gpm—m3/hr
Variable Area (Rotamater)
A
√
√
10:1
M
N
SCFM—Sm3/hr
Vortex Shedding
SD
√
A
√
√
10/1
➅ A
20/5
gpm—m /hr
3
ACFM—Sm3/hr
Fluidic Oscillation (Coanda)
√
A
√
√
12/1
➅ H
20/5
lbm—kgm/hr
Mass Flowmeters Coriolis
√
N
√
√
20:1
M/H
N
SCFM—Sm3/hr
Mass Flowmeters
√
SD
N
√
√
20:1
➆ M
20/5
gpm—m /hr
3
SCFM—Sm3/hr
Thermal Probe
Solids Flowmeters
SD
√
√
5:1 to 80:1
-
5/3
lbm—kgm/hr
gpm—m3/hr
➂
Weirs, Flumes
M
√
SD
100:1
M
4/1
= Non-standard Range
➀ = The data in this column is for general guidance only.
➄ = Varies with upstream disturbance.
L = Limited
➁ = Inherent
rangeability
of primary device
is substantially greater than shown. Value used reflects
➅ = Can be more with high Reynolds number services.
SD = Some Designs
limitations
of
differential
pressure
sensing
device when 1% of rate accuracy is desired. With
➆ = Up to 100:1.
multiple-range intelligent transmitters, rangeability can reach 10:1.
H = High
➇ = More for gas turbine meters.
➂ = Pipe size establishes the upper limit.
A = Average
➈ = Higher and lower flow ranges may be available.
➃ = Practically unlimited with probe type design.
Check several manufacturers.
M = Minimal
N = None
SR = Square Root
14 Volume 4 TRANSACTIONS
1 A Flow Measurement Orientation
TRANSACTIONS Volume 4 15
2
include the use of pitot tubes and fluid. The discharge coefficient k is Re = 3160(SG)(Q)/(ID)m
variable-area meters (rotameters), and influenced by the Reynolds number
are discussed later in this chapter. (see Figure 1-5) and by the “beta where ID is the inside diameter of
ratio,” the ratio between the bore the pipe in inches, Q is the volumet-
Primary Element Options diameter of the flow restriction and ric liquid flow in gallons/minute, SG
In the 18th century, Bernoulli first the inside diameter of the pipe. is the fluid specific gravity at 60°F,
established the relationship between Additional parameters or correc- and m is the viscosity in centipoises.
static and kinetic energy in a flowing tion factors can be used in the deriva- At low Reynolds numbers (gener-
stream. As a fluid passes through a tion of k, depending on the type of ally under Re = 2,000), the flow is
restriction, it accelerates, and the flow element used. These parameters laminar and the velocity profile is
energy for this acceleration is can be computed from equations or parabolic. At high Reynolds num-
obtained from the fluid’s static pres- read from graphs and tables available bers (well over Re = 3,000), the flow
sure. Consequently, the line pressure from the American National becomes fully turbulent, and the
drops at the point of constriction Standards Institute (ANSI), the resulting mixing action produces a
(Figure 2-1). Part of the pressure drop American Petroleum Institute (API), uniform axial velocity across the
is recovered as the flow returns to the the American Society of Mechanical pipe. As shown in Figure 1-5, the
16 Volume 4 TRANSACTIONS
2 Differential Pressure Flowmeters
transition between laminar and tur- find the recommended size, although ential pressure range of 100:1, the
bulent flows can cover a wide range these results should be checked for flowmeter would have an error of
of Reynolds numbers; the relation- reasonableness by hand calculation. ±20% AR. For this reason, differential
ship with the discharge coefficient is producing flowmeters have histori-
a function of the particular primary • Accuracy & Rangeability cally been limited to use within a 3:1
element. The performance of a head-type or 4:1 range.
Today, many engineering societies flowmeter installation is a function Flowmeter rangeability can be fur-
and organizations and most primary of the precision of the flow element ther increased without adverse effect
TRANSACTIONS Volume 4 17
Differential Pressure Flowmeters 2
techniques are cumbersome and compressible fluids, the ratio of dif- inverse derivative algorithm, which
expensive. Intelligent transmitters ferential pressure (h) divided by blocks any rate of change occurring
offer a better option. upstream pressure (P) should not more quickly than the rate at which
The accuracy of intelligent trans- exceed 0.25 (measured in the same the process flow can change.
mitters is usually stated as ±0.1% CS, engineering units).
which includes only errors due to Metering errors due to incorrect • Piping, Installation, & Maintenance
hysteresis, rangeability and linearity. installation of the primary element Installation guidelines are published
Potential errors due to drift, temper- can be substantial (up to 10%). by various professional organizations
ature, humidity, vibration, overrange, Causes of such errors can be the (ISA, ANSI, API, ASME, AGA) and
radio frequency interference and condition of the mating pipe sec- by manufacturers of proprietary
power supply variation are all tions, insufficient straight pipe runs, designs. These guidelines include
excluded. If one includes them, inac- and pressure tap and lead line such recommendations as:
curacy is about 0.2% CS. Because design errors. • When, in addition to measuring
7 Pipe Diameters
Swirl Reducer
Flow
intelligent d/p transmitters can— Under turbulent flow conditions, the flow, the process temperature
based on their own measurements— as much as 10% of the d/p signal can or pressure is also to be measured,
automatically switch ranges between be noise caused by disturbances the pressure transmitter should
two calibrated spans (one for 1-10%, from valves and fittings, both up- and not be installed in the process
the other for 10-100% of FS d/p), it downstream of the element, and by pipe, but should be connected to
should be possible to obtain orifice the element itself. In the majority of the appropriate lead line of the
installations with 1% AR inaccuracy applications, the damping provided flow element via a tee.
over a 10:1 flow range. in d/p cells is sufficient to filter out • Similarly, the thermowell used for
In most flowmetering applications, the noise. Severe noise can be temperature measurement should
density is not measured directly. reduced by the use of two or more be installed at least 10 diameters
Rather, it is assumed to have some pressure taps connected in parallel downstream of the flow element, to
normal value. If density deviates from on both sides of the d/p cell. prevent velocity profile distortions.
this assumed value, error results. Pulsating flow can be caused by • Welds should be ground smooth
Density error can be corrected if it is reciprocating pumps or compressors. and gaskets trimmed so that no
measured directly or indirectly by This pulsation can be reduced by protrusion can be detected by
measuring pressure in gases or temper- moving the flowmeter away from the physical inspection.
ature in liquids. Flow computing pack- source of the pulse, or downstream In order for the velocity profile to
ages are also available that accept the of filters or other dampening fully develop (and the pressure drop
inputs of the d/p transmitter and the devices. Pulsation dampening hard- to be predictable), straight pipe runs
other sensors and can simultaneously ware can also be installed at the are required both up- and down-
calculate mass and volumetric flow. pressure taps, or dampening soft- stream of the d/p element. The
To minimize error (and the need for ware can applied to the d/p cell out- amount of straight run required
density correction) when dealing with put signal. One such filter is the depends on both the beta ratio of
18 Volume 4 TRANSACTIONS
2 Differential Pressure Flowmeters
the installation and on the nature of located as close to the primary ele- out the difference, as long as that
the upstream components in the ment as possible. Lead lines should difference does not change.
pipeline. For example, when a single be as short as possible and of the If the process temperature exceeds
90° elbow precedes an orifice plate, the same diameter. In clean liquid ser- the maximum temperature limitation
straight-pipe requirement ranges from vice, the minimum diameter is G", of the d/p cell, either chemical seals
6 to 20 pipe diameters as the diameter while in condensable vapor service, have to be used or the lead lines need
ratio is increased from 0.2 to 0.8. the minimum diameter is 0.4". In to be long enough to cool the fluid. If
In order to reduce the straight run steam service, the horizontal lead a large temperature drop is required, a
requirement, flow straighteners lines should be kept as short as pos- coiled section of tubing (pigtail) can
(Figure 2-2) such as tube bundles, sible and be tilted (with a minimum be installed in the lead lines to cool
perforated plates, or internal tabs gradient of 1 in/ft with respect to the process fluids.
can be installed upstream of the pri- the piping) towards the tap, so that The frequency of inspection or
mary element. condensate can drain back into the replacement of a primary element
The size and orientation of the pipe. Again, both lead lines should be depends on the erosive and corro-
pressure taps are a function of both exposed to the same ambient condi- sive nature of the process and on the
the pipe size and the type of process tions and be shielded from sunlight. overall accuracy required. If there is
fluid. The recommended maximum In clean liquid or gas service, the lead no previous experience, the orifice
diameter of pressure tap holes lines can be purged through the d/p plate can be removed for inspection
through the pipe or flange is G" for
pipes under 2" in diameter, K" for 2"
and 3" pipes, H" for 4 to 8" and I" for Pipe Taps
larger pipes. Both taps should be of 2 21 D 8D
Flange Taps
the same diameter, and, where the
hole breaks through the inside pipe 1 in. 1 in.
surface, it should be square with no
roughness, burrs, or wire edges.
Connections to pressure holes Flow
should be made by nipples, cou- Corner Taps
plings, or adaptors welded to the
outside surface of the pipe.
On services where the process D D/2
fluid can plug the pressure taps or Figure 2-3: Differential Pressure Tap Location Alternatives
might gel or freeze in the lead lines,
chemical seal protectors can be cell vent or drain connections, and during the first three, six, and 12
used. Connection sizes are usually they should be flushed for several months of its operation. Based on
larger (seal elements can also be minutes to remove all air from the visual inspection of the plate, a rea-
provided with diaphragm exten- lines. Entrapped air can offset the sonable maintenance cycle can be
sions), and, because of the space zero calibration. extrapolated from the findings.
requirement, they are usually Seal pots are on the wet leg in d/p Orifices used for material balance
installed at “radius tap” or “pipe cell installations with small ranges calculations should be on the same
tap” locations, as shown in Figure 2- (under 10 in H2O) in order to mini- maintenance cycle.
3. When chemical seals are used, it mize the level variation in the legs. In
is important that the two connect- steam applications, filling tees are • Sizing the Orifice Plate
ing capillaries, as they are routed to recommended to ensure equal The orifice plate is commonly used
the d/p cell, experience the same height condensate legs on both sides in clean liquid, gas, and steam ser-
temperature and are kept shielded of the d/p cell. If for some reason vice. It is available for all pipe sizes,
from sunlight. the two legs are not of equal height, and if the pressure drop it requires is
The d/p transmitter should be the d/p cell can be biased to zero free, it is very cost-effective for
TRANSACTIONS Volume 4 19
Differential Pressure Flowmeters 2
measuring flows in larger pipes (over process without depressurizing the (Figure 2-3). With corner taps, the
6" diameter). The orifice plate is also line and shutting down flow. In such relatively small clearances represent
approved by many standards organi- fittings, the universal orifice plate, a a potential maintenance problem.
zations for the custody transfer of circular plate with no tab, is used. Vena contracta taps (which are
liquids and gases. The concentric orifice plate close to the radius taps, Figure 2-4)
The orifice flow equations used (Figure 2-4A) has a sharp (square- are located one pipe diameter
today still differ from one another, edged) concentric bore that provides upstream from the plate, and down-
although the various standards orga- an almost pure line contact between stream at the point of vena contrac-
nizations are working to adopt a sin- the plate and the fluid, with negligi- ta. This location varies (with beta
gle, universally accepted orifice flow ble friction drag at the boundary. The ratio and Reynolds number) from
equation. Orifice sizing programs beta (or diameter) ratios of concen- 0.35D to 0.8D.
usually allow the user to select the tric orifice plates range from 0.25 to The vena contracta taps provide
flow equation desired from among 0.75. The maximum velocity and min- the maximum pressure differential,
several. imum static pressure occurs at some but also the most noise. Additionally,
The orifice plate can be made of 0.35 to 0.85 pipe diameters down- if the plate is changed, it may require
any material, although stainless steel stream from the orifice plate. That a change in the tap location. Also, in
is the most common. The thickness point is called the vena contracta. small pipes, the vena contracta might
of the plate used (J-H") is a func- Measuring the differential pressure at lie under a flange. Therefore, vena
tion of the line size, the process tem- a location close to the orifice plate contracta taps normally are used
perature, the pressure, and the differ- minimizes the effect of pipe rough- only in pipe sizes exceeding six inches.
ential pressure. The traditional ori- ness, since friction has an effect on Radius taps are similar to vena
fice is a thin circular plate (with a tab the fluid and the pipe wall. contracta taps, except the down-
for handling and for data), inserted Flange taps are predominantly stream tap is fixed at 0.5D from the
Bevel Where
Vent Hole
Thickness Is
Location
Greater Than
(Liquid
1/8 in (3.175 mm)
Service) or the Orifice
45°
Diameter Is Less
Than 1 in (25 mm)
Flow Orifice
Upstream
Sharp Edge
into the pipeline between the two used in the United States and are orifice plate (Figure 2-3). Pipe taps are
flanges of an orifice union. This located 1 inch from the orifice plate’s located 2.5 pipe diameters upstream
method of installation is cost-effec- surfaces (Figure 2-3). They are not and 8 diameters downstream from
tive, but it calls for a process shut- recommended for use on pipelines the orifice (Figure 2-3). They detect
down whenever the plate is removed under 2 inches in diameter. Corner the smallest pressure difference and,
for maintenance or inspection. In taps are predominant in Europe for because of the tap distance from the
contrast, an orifice fitting allows the all sizes of pipe, and are used in the orifice, the effects of pipe rough-
orifice to be removed from the United States for pipes under 2 inches ness, dimensional inconsistencies,
20 Volume 4 TRANSACTIONS
2 Differential Pressure Flowmeters
and, therefore, measurement errors around the upstream edge of the These plates are usually used in pipe
are the greatest. plate. In extreme cases, this can sizes exceeding four inches in diame-
clog the opening, or it can change ter, and must be carefully installed to
• Orifice Types & Selection the flow pattern, creating measure- make sure that no portion of the
The concentric orifice plate is rec- ment error. Eccentric and segmental flange or gasket interferes with the
ommended for clean liquids, gases, orifice plates are better suited for opening. Flange taps are used with
and steam flows when Reynolds such applications. Concentric ori- both types of plates, and are located
numbers range from 20,000 to 107 in fices are still preferred for multi- in the quadrant opposite the opening
pipes under six inches. Because the phase flows in vertical lines for the eccentric orifice, in line with
basic orifice flow equations assume because accumulation of material is the maximum dam height for the
that flow velocities are well below less likely and the sizing data for segmental orifice.
sonic, a different theoretical and these plates is more reliable. For the measurement of low flow
computational approach is required The eccentric orifice (Figure 2-4B) rates, a d/p cell with an integral
if sonic velocities are expected. The
minimum recommended Reynolds
number for flow through an orifice
(Figure 1-5) varies with the beta ratio
of the orifice and with the pipe size.
In larger size pipes, the minimum
Reynolds number also rises.
Because of this minimum Reynolds
number consideration, square-edged
orifices are seldom used on viscous Flow Flow
fluids. Quadrant-edged and conical
orifice plates (Figure 2-5) are recom- 45°
mended when the Reynolds number
is under 10,000. Flange taps, corner,
and radius taps can all be used with
quadrant-edged orifices, but only
corner taps should be used with a A) Quadrant-Edged B) Conical
conical orifice.
Concentric orifice plates can be Figure 2-5: Orifices for Viscous Flows
provided with drain holes to pre- is similar to the concentric except orifice may be the best choice. In this
vent buildup of entrained liquids in that the opening is offset from the design, the total process flow passes
gas streams, or with vent holes for pipe’s centerline. The opening of the through the d/p cell, eliminating the
venting entrained gases from liquids segmental orifice (Figure 2-4C) is a need for lead lines. These are propri-
(Figure 2-4A). The unmeasured flow segment of a circle. If the secondary etary devices with little published
passing through the vent or drain phase is a gas, the opening of an data on their performance; their flow
hole is usually less than 1% of the eccentric orifice will be located coefficients are based on actual lab-
total flow if the hole diameter is towards the top of the pipe. If the oratory calibrations. They are recom-
less than 10% of the orifice bore. secondary phase is a liquid in a gas or mended for clean, single-phase fluids
The effectiveness of vent/drain a slurry in a liquid stream, the opening only because even small amounts of
holes is limited, however, because should be at the bottom of the pipe. build-up will create significant mea-
they often plug up. The drainage area of the segmental surement errors or will clog the unit.
Concentric orifice plates are not orifice is greater than that of the Restriction orifices are installed to
recommended for multi-phase flu- eccentric orifice, and, therefore, it is remove excess pressure and usually
ids in horizontal lines because the preferred in applications with high operate at sonic velocities with very
secondary phase can build up proportions of the secondary phase. small beta ratios. The pressure drop
TRANSACTIONS Volume 4 21
Differential Pressure Flowmeters 2
across a single restriction orifice process pipe, adequacy of straight laboratory and is provided with honed
should not exceed 500 psid because pipe runs, gasket interference, mis- pipe sections, flow straighteners,
of plugging or galling. In multi-ele- alignment of pipe and orifice bores, senior orifice fittings, and tempera-
ment restriction orifice installations, and lead line design. Other adverse ture controlled enclosures.
the plates are placed approximately conditions include the dulling of the
one pipe diameter from one another sharp edge or nicks caused by corro- • Venturi & Flowtubes
in order to prevent pressure recovery sion or erosion, warpage of the plate Venturi tubes are available in sizes
between the plates. due to waterhammer and dirt, and up to 72", and can pass 25 to 50%
grease or secondary phase deposits more flow than an orifice with the
• Orifice Performance on either orifice surface. Any of the same pressure drop. Furthermore,
Although it is a simple device, the above conditions can change the ori- the total unrecovered head loss
orifice plate is, in principle, a preci- fice discharge coefficient by as much rarely exceeds 10% of measured d/p
sion instrument. Under ideal condi- as 10%. In combination, these prob- (Figure 2-6). The initial cost of ven-
tions, the inaccuracy of an orifice lems can be even more worrisome turi tubes is high, so they are pri-
plate can be in the range of 0.75-1.5% and the net effect unpredictable. marily used on larger flows or on
AR. Orifice plates are, however, quite Therefore, under average operating more difficult or demanding flow
applications. Venturis are insensitive
90
10
to velocity profile effects and
therefore require less straight pipe
run than an orifice. Their contoured
Unrecovered Pressure Loss—Percent of Differential
80
20
Orifice Plate nature, combined with the self-
scouring action of the flow through
70
30
Recovery—Percent of Differential
22 Volume 4 TRANSACTIONS
2 Differential Pressure Flowmeters
length and reduced weight. Pressure Re > 200,000 is between 0.7 and 1.5%. cally coupled to the d/p transmitter
taps are located G to H pipe diame- Flowtubes are often supplied with using filled capillaries. Overall mea-
ter upstream of the inlet cone, and in discharge coefficient graphs because surement accuracy can drop if the
Flow
D d
Throat Outlet
Inlet Inlet
Cone
Cone
the middle of the throat section. the discharge coefficient changes as chemical seal is small, its diaphragm
Piezometer rings can be used with the Reynolds number drops. The is stiff, or if the capillary system is
large venturi tubes to compensate variation in the discharge coefficient not temperature-compensated or
for velocity profile distortions. In of a venturi caused by pipe rough- not shielded from direct sunlight.
slurry service, the pipe taps can be ness is less than 1% because there is
purged or replaced with chemical continuous contact between the • Flow Nozzles
seals, which can eliminate all dead- fluid and the internal pipe surface. The flow nozzle is dimensionally
ended cavities. The high turbulence and the lack of more stable than the orifice plate,
There are several proprietary flow- cavities in which material can accu- particularly in high temperature and
tube designs which provide even mulate make flow tubes well suited high velocity services. It has often
better pressure recovery than the for slurry and sludge services. been used to measure high
classical venturi. The best known of However, maintenance costs can be flowrates of superheated steam.
these proprietary designs is the uni- high if air purging cannot prevent The flow nozzle, like the venturi,
versal venturi (Figure 2-7B). The vari- plugging of the pressure taps and lead has a greater flow capacity than the
ous flowtube designs vary in their lines. Plunger-like devices (vent clean- orifice plate and requires a lower
contours, tap locations, generated ers) can be installed to periodically initial investment than a venturi
d/p and in their unrecovered head
loss. They all have short lay lengths,
typically varying between 2 and 4
Wedge Flow
pipe diameters. These proprietary Element H L
flowtubes usually cost less than the
classical and short-form venturis
because of their short lay length. D
H
However, they may also require more
straight pipe run to condition their
A) Segmental Wedge B) V-Cone
flow velocity profiles.
Flowtube performance is much
Figure 2-8: Proprietary Elements for Difficult Fluids
affected by calibration. The inaccuracy
of the discharge coefficient in a remove buildup from interior open- tube, but also provides less pressure
universal venturi, at Reynolds num- ings, even while the meter is online. recovery (Figure 2-6). A major disad-
bers exceeding 75,000, is 0.5%. The Lead lines can also be replaced with vantage of the nozzle is that it is
inaccuracy of a classical venturi at button-type seal elements hydrauli- more difficult to replace than the
TRANSACTIONS Volume 4 23
Differential Pressure Flowmeters 2
orifice unless it can be removed as accurate way to measure gas flows. dead-ended cavities. The seals attach
part of a spool section. When the gas velocity reaches the to the meter body immediately
The ASME pipe tap flow nozzle is speed of sound in the throat, the upstream and downstream of the
predominant in the United States velocity cannot increase any more restriction. They rarely require clean-
(Figure 2-7C). The downstream end (even if downstream pressure is ing, even in services like dewatered
of a nozzle is a short tube having the reduced), and a choked flow condi- sludge, black liquor, coal slurry, fly
same diameter as the vena contrac- tion is reached. Such “critical flow ash slurry, taconite, and crude oil.
ta of an equivalent orifice plate. The nozzles” are very accurate and often The minimum Reynolds number is
low-beta designs range in diameter are used in flow laboratories as stan- only 500, and the meter requires
ratios from 0.2 to 0.5, while the high dards for calibrating other gas only five diameters of upstream
beta-ratio designs vary between flowmetering devices. straight pipe run.
0.45 and 0.8. The nozzle should Nozzles can be installed in any The segmental wedge has a
always be centered in the pipe, and position, although horizontal orien- V-shaped restriction characterized
the downstream pressure tap tation is preferred. Vertical down- by the H/D ratio, where H is the
should be inside the nozzle exit. The flow is preferred for wet steam, height of the opening below the
throat taper should always decrease gases, or liquids containing solids. restriction and D is the diameter. The
the diameter toward the exit. Flow The straight pipe run requirements H/D ratio can be varied to match the
nozzles are not recommended for are similar to those of orifice plates. flow range and to produce the
slurries or dirty fluids. The most desired d/p. The oncoming flow cre-
common flow nozzle is the flange • Segmental Wedge Elements ates a sweeping action through the
type. Taps are commonly located The segmental wedge element (Figure meter. This provides a scouring effect
one pipe diameter upstream and H 2-8A) is a proprietary device designed on both faces of the restriction,
pipe diameter downstream from for use in slurry, corrosive, erosive, helping to keep it clean and free of
the inlet face. viscous, or high-temperature applica- buildup. Segmental wedges can mea-
Flow nozzle accuracy is typically tions. It is relatively expensive and is sure flow in both directions, but the
d/p transmitter must be calibrated
Impact Pressure Connection for a split range, or the flow element
must be provided with two sets of
connections for two d/p transmit-
Stainless Steel Tubing
Tubing Adaptor ters (one for forward and one for
reverse flow).
P Pt An uncalibrated wedge element
Static Pressure
Static Pressure Holes
Connection can be expected to have a 2% to 5%
Outer Pipe Only
(P) AR inaccuracy over a 3:1 range. A cal-
Vp Vp ~ Pt - P ibrated wedge element can reduce
that to 0.5% AR if the fluid density is
Impact Pressure Opening (Pt) constant. If slurry density is variable
Figure 2-9: Pitot Tubes Measure Two Pressures and/or unmeasured, error rises.
1% AR, with a potential for 0.25% AR used mostly on difficult fluids, where • Venturi-Cone Element
if calibrated. While discharge coeffi- the dramatic savings in maintenance The venturi-cone (V-cone) element
cient data is available for Reynolds can justify the initial cost. The unique (Figure 2-8B) is another proprietary
numbers as low as 5,000, it is advis- flow restriction is designed to last the design that promises consistent per-
able to use flow nozzles only when life of the installation without deteri- formance at low Reynolds numbers
the Reynolds number exceeds 50,000. oration. and is insensitive to velocity profile
Flow nozzles maintain their accuracy Wedge elements are used with distortion or swirl effects. Again, how-
for long periods, even in difficult ser- 3-in diameter chemical seals, elimi- ever, it is relatively expensive. The V-
vice. Flow nozzles can be a highly nating both the lead lines and any cone restriction has a unique geometry
24 Volume 4 TRANSACTIONS
2 Differential Pressure Flowmeters
TRANSACTIONS Volume 4 25
Differential Pressure Flowmeters 2
26 Volume 4 TRANSACTIONS
2 Differential Pressure Flowmeters
TRANSACTIONS Volume 4 27
Differential Pressure Flowmeters 2
an indication of the average flow in same advantages and disadvantages the manifolds can be purged to keep
the pipe (Figure 2-12). The port closer as do single-port tubes. They are the ports clean.
to the outlet of the combined signal slightly more expensive and a little Because area-averaging pitot sta-
has a slightly greater influence, than more accurate, especially if the flow tions generate very small pressure dif-
the port that is farthest away, but, for is not fully formed. Some averaging ferentials, it may be necessary to use
secondary applications where pitot pitot sensors can be inserted through low differential d/p cells with spans
tubes are commonly used, this error the same opening (or hot tap) which as low as 0-0.01 in water column. To
is acceptable. accommodates a single-port tube. improve accuracy, a hexagonal cell-
The number of impact ports, the type flow straightener and a flow
distance between ports, and the • Area Averaging nozzle can be installed upstream of
diameter of the averaging pitot tube Area-averaging pitot stations are the area-averaging pitot flow sensor.
all can be modified to match the used to measure the large flows of The flow straightener removes local
needs of a particular application. low pressure air in boilers, dryers, or turbulence, while the nozzle ampli-
Sensing ports in averaging pitot tubes HVAC systems. These units are avail- fies the differential pressure pro-
are often too large to allow the tube able for the various standard sizes of duced by the sensor.
to behave as a true averaging cham- circular or rectangular ducts (Figure
ber. This is because the oversized 2-13) and for pipes. They are so • Installation
Pitot tubes can be used as permanently
installed flow sensors or as portable
Installed
monitoring devices providing periodic
Drill
data. Permanently installed carbon
Thru
steel or stainless steel units can oper-
Valve
ate at up to 1400 PSIG pressures and
Inserted are inserted into the pipe through
flanged or screw connections. Their
installation usually occurs prior to
plant start-up, but they can be hot-
tapped into an operating process.
In a hot-tap installation (Figure
2-14), one first welds a fitting to the
pipe. Then a drill-through valve is
attached to the fitting and a hole is
drilled through the pipe. Then, after
partially withdrawing the drill, the
valve is closed, the drill is removed
and the pitot tube is inserted. Finally,
the valve is opened and the pitot
tube is fully inserted.
The velocity profile of the flowing
stream inside the pipe is affected by
Figure 2-14: Hot Tap Installation of a Pitot Tube the Reynolds number of the flowing
port openings are optimized not for designed that each segment of the fluid, pipe surface roughness, and by
averaging, but to prevent plugging. In cross-section is provided with both upstream disturbances, such as
some installations, purging with an an impact and a static pressure port. valves, elbows, and other fittings.
inert gas is used to keep the ports Each set of ports is connected to its Pitot tubes should be used only if the
clean, allowing the sensor to use own manifold, which combines the minimum Reynolds number exceeds
smaller ports. average static and average impact 20,000 and if either a straight run of
Averaging pitot tubes offer the pressure signals. If plugging is likely, about 25 diameters can be provided
28 Volume 4 TRANSACTIONS
2 Differential Pressure Flowmeters
upstream to the pitot tube or if plus: 1.25 in for K-in diameter probes; spring is used to return the flow ele-
straightening vanes can be installed. 1.5 in for H-in; 1.56 in for I-in; and ment to its resting position when the
1.94 in for 1-in diameter probes. flow lessens. Gravity-operated meters
• Vibration Damage Once the velocity limits have been (rotameters) must be installed in a ver-
Natural frequency resonant vibra- calculated, make sure that they do tical position, whereas spring operated
tions can cause pitot tube failure. not fall within the range of operating ones can be mounted in any position.
100
90
80
Gravity
70
Equilibrium
60
50
Float
40
30
Flow
20
10
Piston in
Scale
Tapered Tube
Perforated
Flexing Vane,
Tapered Metering Tube (Rotameter) Tapered Plug Cylinder Disc, or Flapper
Natural frequency vibration is caused velocities. If they do, change the All variable area flowmeters are avail-
by forces created as vortices are shed probe diameter, or its mounting, or able with local indicators. Most can
by the pitot tube. The pitot tube is do both, until there is no overlap. also be provided with position sensors
expected to experience such vibra- and transmitters (pneumatic, electronic,
tion if the process fluid velocity (in Variable Area Flowmeters digital, or fiberoptic) for connecting to
feet per second) is between a lower Variable area flowmeters (Figure 2-15) remote displays or controls.
limit (VL) and an upper limit (VH). The are simple and versatile devices that
values of VL and VH can be calculated operate at a relatively constant pres- • Purge-Flow Regulators
(for the products of a given manufac- sure drop and measure the flow of liq- If a needle valve is placed at the
turer) using the equations below. uids, gases, and steam. The position of inlet or outlet of a rotameter, and a
their float, piston or vane is changed d/p regulator controls the pressure
VL = 5253(M x Pr x D)/L2 as the increasing flow rate opens a difference across this combination,
VH = 7879(M x Pr x D)/L2 larger flow area to pass the flowing the result is a purge-flow regulator.
fluid. The position of the float, piston Such instrumentation packages are
Where M = mounting factor (3.52 for or vane provides a direct visual indica- used as self-contained purge
single mount); Pr = probe factor (0.185 tion of flow rate. Design variations flowmeters (Figure 2-16). These are
for K-in diameter probes; 0.269 for include the rotameter (a float in a among the least expensive and most
H-in; 0.372 for I-in; and 0.552 for 1-in); tapered tube), orifice/rotameter widely used flowmeters. Their main
D = probe diameter (inches); L = combination (bypass rotameter), application is to control small gas or
unsupported probe length in inches, open-channel variable gate, tapered liquid purge streams. They are used
which is calculated as the sum of the plug, and vane or piston designs. to protect instruments from con-
pipe I.D. plus the pipe wall thickness Either the force of gravity or a tacting hot and corrosive fluids, to
TRANSACTIONS Volume 4 29
Differential Pressure Flowmeters 2
protect pressure taps from plugging, and the inlet pressure (P1) is variable. combination of the buoyancy of the
to protect the cleanliness of optical They can handle extremely small liquid and the velocity head of the
devices, and to protect electrical flow rates from 0.01 cc/min for liq- fluid. With gases, buoyancy is negligi-
devices from igniting upon contact uids and from 0.5 cc/min for gases. ble, and the float responds mostly to
with combustibles. The most common size is a glass the velocity head.
Purge meters are quite useful in tube rotameter with G-in (6 mm) In a rotameter (Figure 2-15), the
adding nitrogen gas to the vapor connections, a range of 0.05-0.5 gpm metering tube is mounted vertically,
with the small end at the bottom. The
fluid to be measured enters at the
Flow at
bottom of the tube, passes upward
P0 Outlet Pressure around the float, and exits the top.
When no flow exists, the float rests at
the bottom. When fluid enters, the
metering float begins to rise.
The float moves up and down in
proportion to the fluid flow rate and
Tube the annular area between the float
and the tube wall. As the float rises,
Spring #1
Float the size of the annular opening
Diaphragm increases. As this area increases, the
differential pressure across the float
P2 decreases. The float reaches a stable
position when the upward force
Regulator
exerted by the flowing fluid equals
Valve Flow Control
Valve (V) the weight of the float. Every float
Spring #2 position corresponds to a particular
flowrate for a particular fluid’s densi-
Flow at
P1 Inlet Pressure ty and viscosity. For this reason, it is
necessary to size the rotameter for
Figure 2-16: Purge Flowmeter Design each application. When sized cor-
rectly, the flow rate can be deter-
spaces of tanks and other equip- (0.2-2.0 lpm) on water or 0.2-2.0 scfm mined by matching the float position
ment. Purging with nitrogen gas (0.3-3.0 cmph) in air service. Typical to a calibrated scale on the outside
reduces the possibility of developing accuracy is ±5% FS over a 10:1 range, of the rotameter. Many rotameters
a flammable mixture because it dis- and the most common pressure rat- come with a built-in valve for adjust-
places flammable gases. The purge- ing is 150 psig (1 MPa). ing flow manually.
flow regulator is reliable, intrinsically Several shapes of float are avail-
safe, and inexpensive. • Rotameters able for various applications. One
As shown in Figure 2-16, purge The rotameter is the most widely early design had slots, which caused
meters can operate in the constant used variable area flowmeter the float to spin for stabilizing and
flow mode, where P2 - P0 is held con- because of its low cost, simplicity, centering purposes. Because this
stant at about 60 to 80 in H2O low pressure drop, relatively wide float rotated, the term rotameter
differential. In bubbler and purge rangeability, and linear output. Its was coined.
applications, the inlet pressure (P1) is operation is simple: in order to pass Rotameters are typically provided
held constant and the outlet pres- through the tapered tube, the fluid with calibration data and a direct
sure (P0) is variable. Figure 2-16 flow raises the float. The greater the reading scale for air or water (or
describes a configuration where the flow, the higher the float is lifted. In both). To size a rotameter for other
outlet pressure (P0) is held constant liquid service, the float rises due to a service, one must first convert the
30 Volume 4 TRANSACTIONS
2 Differential Pressure Flowmeters
actual flow to a standard flow. For liq- tic. Some fluids attack the glass Glass-tube rotameters are often
uids, this standard flow is the water metering tube, such as wet steam or used in applications where several
equivalent in gpm; for gases, the stan- high-pH water over 194°F (which can streams of gases or liquids are being
dard flow is the air flow equivalent in soften glass); caustic soda (which dis- metered at the same time or mixed in
standard cubic feet per minute (scfm). solves glass); and hydrofluoric acid a manifold, or where a single fluid is
Tables listing standard water equiva- (which etches glass). being exhausted through several
lent gpm and/or air scfm values are Floats have a sharp edge at the channels (Figure 2-17). Multiple tube
provided by rotameter manufacturers. point where the reading should be flowmeters allow up to six rotameters
Manufacturers also often provide observed on the tube-mounted to be mounted in the same frame.
slide rules, nomographs, or computer scale. For improved reading accuracy, It also is possible to operate a
software for rotameter sizing. a glass-tube rotameter should be
installed at eye level. The scale can
• Design Variations be calibrated for direct reading of air
A wide choice of materials is available or water, or can read percentage of
for floats, packing, O-rings, and end range. In general, glass tube rotame-
fittings. Rotameter tubes for such ters can measure flows up to about
safe applications as air or water can 60 gpm water and 200 scfh air.
be made of glass, whereas if breakage A correlation rotameter has a
would create an unsafe condition, scale from which a reading is taken
they are provided with metal tubes. (Figure 2-15). This reading is then
Glass tubes are most common, being compared to a correlation table for a
precision formed of safety shielded given gas or liquid to get the actual
flow in engineering units. Correlation
Figure 2-17: Multi-Tube Rotameter Station
charts are readily available for nitro-
gen, oxygen, hydrogen, helium, argon, rotameter in a vacuum. If the
and carbon dioxide. While not nearly rotameter has a valve, it must be
as convenient as a direct reading placed at the outlet at the top of the
device, a correlation meter is more meter. For applications requiring a
accurate. This is because a direct- wide measurement range, a dual-ball
reading device is accurate for only rotameter can be used. This instru-
one specific gas or liquid at a partic- ment has two ball floats: a light ball
ular temperature and pressure. A cor- (typically black) for indicating low
relation flowmeter can be used with flows and a heavy ball (usually white)
a wide variety of fluids and gases for indicating high flows. The black
under various conditions. In the same ball is read until it goes off scale, and
tube, different flow rates can be han- then the white ball is read. One such
dled by using different floats. instrument has a black measuring
Small glass tube rotameters are suit- range from 235-2,350 ml/min and a
able for working with pressures up to white to 5,000 ml/min.
Rotameters can be specified in a wide range of 500 psig, but the maximum operating For higher pressures and tempera-
sizes and materials. pressure of a large (2-in diameter) tube tures beyond the practical range of
may be as low as 100 psig. The practi- glass, metal tube rotameters can be
borosilicate glass. Floats typically are cal temperature limit is about 400°F, used. These tubes are usually made
machined from glass, plastic, metal, but such high-temperature operation of stainless steel, and the position of
or stainless steel for corrosion resis- substantially reduces the operating the float is detected by magnetic fol-
tance. Other float materials include pressure of the tube. In general, there lowers with readouts outside the
carboloy, sapphire, and tantalum. End is a linear relationship between oper- metering tube.
fittings are available in metal or plas- ating temperature and pressure. Metal-tube rotameters can be
TRANSACTIONS Volume 4 31
Differential Pressure Flowmeters 2
32 Volume 4 TRANSACTIONS
2 Differential Pressure Flowmeters
sizes and the requirement that it be One gate-type variable-area tion. The inaccuracy of such meters
installed vertically (there may not be flow-meter resembles a butterfly is 2-5% FS. The meter can be used
enough head room). The cost of a valve. Flow through the meter on oil, water and air, and is available
large rotameter installation can be forces a spring-loaded vane to in sizes up to 4 inches. It also is used
reduced by using an orifice bypass or rotate, and a mechanical connec- as an indicating flow switch in safe-
a pitot tube in combination with a tion provides local flow rate indica- ty interlock systems. T
smaller rotameter. The same-size
bypass rotameter can be used to
measure a variety of flows, with the References & Further Reading
only difference between applications • OMEGA Complete Flow and Level Measurement Handbook and
being the orifice plate and the differ- Encyclopedia®, OMEGA Press, 1995.
ential it produces. • OMEGA Volume 29 Handbook & Encyclopedia, Purchasing Agents
Advantages of a bypass rotameter Edition, OMEGA Press, 1995.
include low cost; its major disadvan- • “Choices Abound in Flow Measurement”, D. Ginesi, Chemical Engineering,
tage is inaccuracy and sensitivity to April 1991.
material build-up. Bypass rotameters • “Developments in DP Flowmeters,” Jesse Yoder, Control, April 1998.
are often provided with isolation • Differential Producers - Orifice, Nozzle, Venturi, ANSI/ASME MFC,
valves so that they can be removed December 1983.
for maintenance without shutting • Flow Measurement Engineers’ Handbook, R.W. Miller, McGraw-Hill, 1996.
down the process line. • Flow Measurement, D.W. Spitzer, Instrument Society of America, 1991.
Tapered plug flowmeters are vari- • Flow of Water Through Orifices, AGA/ASME, Ohio State Univ. Bulletin
able-area flowmeters with a station- 89, Vol. IV, No. 3.
ary core and a piston that moves as • Fluid Meters, H.S. Bean , American Society of Mechanical Engineers, 1971.
the flow varies. In one design, the • Fundamentals of Flow Measurement, J. P. DeCarlo, Instrument Society of
piston movement mechanically America, 1984.
moves a pointer, while in another it • Instrument Engineers Handbook, 3rd edition, Bela Liptak, CRC Press, 1995.
magnetically moves an external • “Orifice Metering of Natural Gas”, AGA Report 3, 1985.
flow rate indicator. The second • “Primary Element Solves Difficult Flow Metering Problems at Water
design has a metallic meter body for Waste Treatment Plant,” D. Ginesi, L. Keefe, and P. Miller, Proceedings of
applications up to 1,000 psig. ISA 1989, Instrument Society of America, 1989.
TRANSACTIONS Volume 4 33
3
Disc Vane
Housing Vane Slot
Ball
Rotor
while also adding kinetic energy to with turndowns as high as 100:1, pressure drop through the meter also
the fluid. although ranges of 15:1 or lower are rises. Consequently, the maximum
The turbine flowmeter consists of a much more common. Slippage (and minimum) flow capacity of the
multi-bladed rotor mounted at right between the flowmeter components flowmeter is decreased as viscosity
34 Volume 4 TRANSACTIONS
3 Mechanical Flowmeters
increases. The higher the viscosity, Because it must be nonmagnetic, of these meters is required to be
the less slippage and the lower the the meter housing is usually made of ±2% of actual flow rate. Higher vis-
measurable flow rate becomes. As bronze but can be made from plastic cosity can produce higher accuracy,
viscosity decreases, the low flow for corrosion resistance or cost while lower viscosity and wear over
Measuring Chamber
Cylindrical
Abutment Control Roller
Slide
Piston
Valve
Piston
Hub
Housing
Piston
Measuring
Chamber Inlet
performance of the meter deterio- savings. The wetted parts such as the time will reduce accuracy. The AWWA
rates. The maximum allowable pres- disc and spindle are usually bronze, requires that residential water meters
sure drop across the meter con- rubber, aluminum, neoprene, Buna-N, be re-calibrated every 10 years.
strains the maximum operating flow or a fluoroelastomer such as Viton®. Because of the intermittent use
in high viscosity services. Nutating disc meters are designed patterns of residential users, this cor-
for water service and the materials of responds to recalibrating L x I in
• Liquid PD Meters which they are made must be residential water meters after they
Nutating disc meters are the most checked for compatibility with other have metered 5 million gallons. In
common PD meters. They are used as fluids. Meters with rubber discs give industrial applications, however, these
residential water meters around the better accuracy than metal discs due meters are likely to pass this threshold
world. As water flows through the to the better sealing they provide. much sooner. The maximum continu-
metering chamber, it causes a disc to Nutating disc meters are available ous flow of a nutating disc meter is
wobble (nutate), turning a spindle, in L-in to 2-in sizes. They are suited usually about 60-80% of the maxi-
which rotates a magnet. This magnet for 150-psig operating pressures with mum flow in intermittent service.
is coupled to a mechanical register overpressure to a maximum of 300 Rotating vane meters (Figure 3-1B)
or a pulse transmitter. Because the psig. Cold water service units are have spring-loaded vanes that entrap
flowmeter entraps a fixed quantity temperature-limited to 120°F. Hot increments of liquid between the
of fluid each time the spindle is water units are available up to 250°F. eccentrically mounted rotor and the
rotated, the rate of flow is propor- These meters must meet American casing. The rotation of the vanes
tional to the rotational velocity of Water Works Association (AWWA) moves the flow increment from inlet
the spindle (Figure 3-1A). standards for accuracy. The accuracy to outlet and discharge. Accuracy of
TRANSACTIONS Volume 4 35
Mechanical Flowmeters 3
±0.1% of actual rate (AR) is normal, large particle size or abrasive solids. are rated to 1,500 psig. They can meter
and larger size meters on higher vis- The measurement chamber is flow rates from 1 gpm to 65 gpm in
cosity services can achieve accuracy cylindrical with a partition plate sep- continuous service with intermittent
to within 0.05% of rate. arating its inlet port from its outlet. excursions to 100 gpm. Meters are
Rotating vane meters are regularly The piston is also cylindrical and is sized so that pressure drop is below
used in the petroleum industry and punctured by numerous openings to 35 psid at maximum flow rate.
are capable of metering solids-laden allow free flow on both sides of the Accuracy ranges from ±0.5 % AR for
crude oils at flow rates as high as piston and the post (Figure 3-2A). The viscous fluids to ±2% AR for nonvis-
17,500 gpm. Pressure and temperature piston is guided by a control roller cous applications. Upper limit on
limits depend on the materials of within the measuring chamber, and viscosity is 10,000 centipoise.
construction and can be as high as the motion of the piston is trans- Reciprocating piston meters are
350°F and 1,000 psig. Viscosity limits ferred to a follower magnet which is probably the oldest PD meter designs.
are 1 to 25,000 centipoise. external to the flowstream. The fol- They are available with multiple pis-
In the rotary displacement meter, lower magnet can be used to drive tons, double-acting pistons, or rotary
a fluted central rotor operates in either a transmitter, a register, or pistons. As in a reciprocating piston
constant relationship with two wiper both. The motion of the piston is engine, fluid is drawn into one piston
rotors in a six-phase cycle. Its appli- oscillatory (not rotary) since it is con- chamber as it is discharged from the
cations and features are similar to strained to move in one plane. The opposed piston in the meter.
those of the rotary vane meter. rate of flow is proportional to the Typically, either a crankshaft or a hor-
rate of oscillation of the piston. izontal slide is used to control the
• Piston Meters The internals of this flowmeter can opening and closing of the proper ori-
Oscillating piston flowmeters typical- be removed without disconnection of fices in the meter. These meters are
ly are used in viscous fluid services the meter from the pipeline. Because usually smaller (available in sizes
such as oil metering on engine test of the close tolerances required to down to 1/10-in diameter) and are
stands where turndown is not critical seal the piston and to reduce slippage, used for measuring very low flows of
(Figure 3-2). These meters also can be these meters require regular mainte- viscous liquids.
used on residential water service and nance. Oscillating piston flow meters
can pass limited quantities of dirt, are available in H-in to 3-in sizes, and • Gear & Lobe Meters
such as pipe scale and fine (viz,-200 can generally be used between 100 The oval gear PD meter uses two
mesh or -74 micron) sand, but not and 150 psig. Some industrial versions fine-toothed gears, one mounted
36 Volume 4 TRANSACTIONS
3 Mechanical Flowmeters
horizontally, the other vertically, at high flows and can be used at high rotating impellers is sensed by prox-
with gears meshing at the tip of the operating pressures (to 1,200 psig) imity switches (usually Hall-effect
vertical gear and the center of the and temperatures (to 400°F). detectors) mounted external to the
horizontal gear (Figure 3-3A). The two The lobe gear meter is available in flow chamber. The sensor transmits a
rotors rotate opposite to each other, a wide range of materials of con- pulse train to a counter or flow con-
creating an entrapment in the cres- struction, from thermoplastics to troller. These meters are available in
cent-shaped gap between the hous- highly corrosion-resistant metals. 1/10-in to 6-in sizes and can handle
ing and the gear. These meters can be Disadvantages of this design include a pressures to 3,000 psig and tempera-
very accurate if slippage between the loss of accuracy at low flows. Also, tures to 400°F.
housing and the gears is kept small. If the maximum flow through this meter
the process fluid viscosity is greater is less than for the same size oscillato- • Helix Meters
than 10 centipoise and the flowrate is ry piston or nutating disc meter. The helix meter is a positive dis-
above 20% of rated capacity, accura- In the rotating impeller meter, placement device that uses two radi-
cy of 0.1% AR can be obtained. At very coarse gears entrap the fluid ally pitched helical gears to continu-
lower flows and at lower viscosity, and pass a fixed volume of fluid ously entrap the process fluid as it
slippage increases and accuracy with each rotation (Figure 3-3C). flows. The flow forces the helical
decreases to 0.5% AR or less. These meters are accurate to 0.5% gears to rotate in the plane of the
The lubricating characteristics of of rate if the viscosity of the pipeline. Optical or magnetic sensors
the process fluid also affect the turn- process fluid is both high and con- are used to encode a pulse train pro-
down of an oval gear meter. With liq- stant, or varies only within a narrow portional to the rotational speed of
uids that do not lubricate well, maxi- band. These meters can be made the helical gears. The forces required
mum rotor speed must be derated to out of a variety of metals, including to make the helices rotate are rela-
limit wear. Another way to limit wear stainless steel, and corrosion-resis- tively small and therefore, in com-
is to keep the pressure drop across tant plastics such as PVDF (Kynar). parison to other PD meters, the
the meter below 15 psid. Therefore, These meters are used to meter pressure drop is relatively low. The
the pressure drop across the meter paints and, because they are avail- best attainable accuracy is about
limits the allowable maximum flow able in 3A or sanitary designs, also ±0.2% or rate.
in high viscosity service.
Rotating lobe and impeller type
PD meters are variations of the oval
+10
gear flowmeter that do not share its
precise gearing. In the rotating lobe
design, two impellers rotate in oppo- +1.0
site directions within the ovoid
housing (Figure 3-3B). As they rotate,
Error
0.1
>1000cP
a fixed volume of liquid is entrapped 300cP
and then transported toward the 100cP
-1.0
outlet. Because the lobe gears
remain in a fixed relative position, it 30cP 10cP 3cP
is only necessary to measure the -10
rotational velocity of one of them. 0.1 1.0 10 100
The impeller is either geared to a reg- Maximum Rated Flow, %
ister or is magnetically coupled to a
transmitter. Lobe meters can be fur- Figure 3-4: Effect of Viscosity on Low-Flow Accuracy
nished in 2-in to 24-in line sizes. Flow
capacity is 8-10 gpm to 18,000 gpm in milk, juices, and chocolate. As shown in Figure 3-4, measure-
the larger sizes. They provide good In these units, the passage of mag- ment error rises as either the operat-
repeatability (better than 0.015% AR) nets embedded in the lobes of the ing flowrate or the viscosity of the
TRANSACTIONS Volume 4 37
Mechanical Flowmeters 3
process fluid drops. Helical gear which also serves to position the tub- and the required flow rate and dis-
meters can measure the flow of highly ing. This type of metering pump is used charge pressure. Check valves (or, on
viscous fluids (from 3 to 300,000 cP), in laboratories, in a variety of medical critical applications, double check
making them ideal for extremely applications, in the majority of envi- valves) are selected to protect
Discharge Discharge
Adjustable
Outlet Valve
Diaphragm
Piston
Piston
Support
Packing Gland
Plates
(Stuffing Box)
Hydraulic
Suction
Oil Inlet Valve
Check Valve
thick fluids such as glues and very ronmental sampling systems, and also against backflow.
viscous polymers. Because at maxi- in dispensing hypochlorite solutions. Diaphragm pumps are the most
mum flow the pressure drop through The tubing can be silicone-rubber or, if common industrial PD pumps (Figure
the meter should not exceed 30 psid, a more corrosion-resistant material is 3-5B). A typical configuration consists
the maximum rated flow through the desired, PTFE tubing. of a single diaphragm, a chamber, and
meter is reduced as the fluid viscosi- Piston pumps deliver a fixed vol- suction and discharge check valves
ty increases. If the process fluid has ume of liquid with each “out” stroke to prevent backflow. The piston can
good lubricating characteristics, the and a fixed volume enters the cham- either be directly coupled to the
meter turndown can be as high as ber on each “in” stroke (Figure 3-5A). diaphragm or can force a hydraulic
100:1, but lower (10:1) turndowns are Check valves keep the fluid flow oil to drive the diaphragm. Maximum
more typical. from reversing. As with all positive output pressure is about 125 psig.
displacement pumps, piston pumps Variations include bellows-type
• Metering Pumps generate a pulsating flow. To mini- diaphragms, hydraulically actuated
Metering pumps are PD meters that mize the pulsation, multiple pistons double diaphragms, and air-operat-
also impart kinetic energy to the or pulsation-dampening reservoirs ed, reciprocating double-diaphragms.
process fluid. There are three basic are installed. Because of the close
designs: peristaltic, piston, and tolerances of the piston and cylinder • Gas PD Meters
diaphragm. sleeve, a flushing mechanism must be PD gas meters operate by counting
Peristaltic pumps operate by having provided in abrasive applications. the number of entrapped volumes
fingers or a cam systematically squeeze Piston pumps are sized on the basis of gas passed, similar to the way PD
a plastic tubing against the housing, of the displacement of the piston meters operate on liquids. The
38 Volume 4 TRANSACTIONS
3 Mechanical Flowmeters
primary difference is that gases are calibrated for natural gas, which has a operate with little or no maintenance
compressible. specific gravity of 0.6 (relative to air). indefinitely.
Diaphragm gas meters most often Therefore, it is necessary to re-cali- Lobe gear meters (or lobed
are used to measure the flow of nat- brate the flow rating of the meter impeller meters, as they are also
ural gas, especially in metering con- when it is used to meter other gases. known), also are used for gas service.
sumption by households. The meter The calibration for the new flow rat- Accuracy in gas service is ±1% of rate
is constructed from aluminum cast- ing (QN) is obtained by multiplying over a 10:1 turndown, and typical
ings with cloth-backed rubber the meter’s flow rating for natural gas pressure drop is 0.1 psid. Because of
diaphragms. The meter consists of (QC) by the square root of the ratio the close tolerances, upstream filtra-
four chambers: the two diaphragm of the specific gravities of natural gas tion is required for dirty lines.
chambers on the inlet and outlet (0.6) and the new gas (SGN): Rotating vane meters measure the
sides and the inlet and outlet cham- flow of gas in the same ranges as do
bers of the meter body. The passage QN= QC(0.6/SGN)0.5 lobe gear meters (up to 100,000
of gas through the meter creates a ft3/hr) but can be used over a wider
differential pressure between the two Diaphragm meters are usually rated 25:1 turndown. They also incur a lower
diaphragm chambers by compressing in units of cubic feet per hour and pressure drop of 0.05 in H2O for sim-
the one on the inlet side and expand- sized for a pressure drop of 0.5-2 in ilar accuracy, and, because the clear-
ing the one on the outlet side. This H2O. Accuracy is roughly ±1% of read- ances are somewhat more forgiving,
action alternately empties and fills ing over a 200:1 range. They maintain upstream filtration is not as critical.
the four chambers. The slide valves at their accuracy for long periods of
the top of the meter alternate the time, which makes them good choices • High-Precision PD Systems
roles of the chambers and synchro- for retail revenue metering applica- High-precision gas meters are usually
nize the action of the diaphragms, as tions. Unless the gas is unusually dirty a hybrid combining a standard PD
Displacement
Displacement
DC Motor Flowmeter
Flowmeter
M
Gas Flow M
PDC
Low
Sensitivity
Leaf
Displacement
Transducers
PDC
High
Sensitivity
Leaf
Differential
Pressure
Zeroing Solenoids Detection Piston
A) Gas Service B) Liquid Service
well as operating the crank mecha- (producer gas, or recycled methane meter and a motor drive that elimi-
nism for the meter register. from composting or digesting, for nates the pressure drop across the
Diaphragm meters generally are example), the diaphragm meter will meter. Equalizing the inlet and outlet
TRANSACTIONS Volume 4 39
Mechanical Flowmeters 3
pressures eliminates slip flows, leak- Vapor separators are usually includ- Therefore, field-mounted and in-line
age, and blow-by. In high-precision ed, to prevent vapor lock. provers have been developed. This
gas flowmeter installations, high- type of prover consists of a calibrat-
sensitivity leaves are used to detect • Testing, Calibration & Provers ed chamber equipped with a barrier
the pressure differential, and dis- All meters with moving parts require piston (Figure 3-7). Two detectors are
placement transducers are used to periodic testing, recalibration and mounted a known distance (and
measure the deflection of the leaves repair, because wear increases the therefore a known volume) apart. As
(Figure 3-6A). Designed to operate at clearances. Recalibration can be the flow passes through the cham-
ber, the displacer piston is moved
downstream. Dividing the volume of
the chamber by the time it takes for
1st Detector Flow Tube 2nd Detector the displacer to move from one
detector to the other gives the cali-
brated flow rate. This rate is then
compared to the reading of the
Displacer
40 Volume 4 TRANSACTIONS
3 Mechanical Flowmeters
of a multi-bladed rotor mounted at are made of permanently magnetized they are in the presence of a very low
right angles to the flow and suspend- material (Figure 3-8B). As each blade strength (on the order of 25 gauss)
ed in the fluid stream on a free-run- passes the coil, it generates a voltage magnetic field.
ning bearing. The diameter of the pulse. In some designs, only one blade In these turbine flowmeters, very
rotor is very slightly less than the is magnetic and the pulse represents a small magnets are embedded in the
inside diameter of the metering complete revolution of the rotor. tips of the rotor blades. Rotors are typ-
chamber, and its speed of rotation is The outputs of reluctance and ically made of a non-magnetic materi-
proportional to the volumetric flow inductive pick-up coils are continu- al, like polypropylene, Ryton, or PVDF
rate. Turbine rotation can be detect- ous sine waves with the pulse train’s (Kynar). The signal output from a Hall-
ed by solid state devices (reluctance, frequency proportional to the flow effect sensor is a square wave pulse
inductance, capacitive and Hall- rate. At low flow, the output (the train, at a frequency proportional to
effect pick-ups) or by mechanical height of the voltage pulse) may be the volumetric flowrate.
sensors (gear or magnetic drives). on the order of 20 mV peak-to-peak. Because Hall-effect sensors have no
In the reluctance pick-up, the coil It is not advisable to transport such a magnetic drag, they can operate at
is a permanent magnet and the tur- weak signal over long distances. lower flow velocities (0.2 ft/sec) than
bine blades are made of a material Therefore, the distance between the magnetic pick-up designs (0.5-1.0
attracted to magnets. As each blade pickup and associated display elec- ft/sec). In addition, the Hall-effect sen-
passes the coil, a voltage is generated tronics or preamplifier must be short. sor provides a signal of high amplitude
Permanent
Magnet Coil
Cone
Coil
Blade
Rotor
N Permanent
Magnet
Reluctance Pickup Coil
S
in the coil (Figure 3-8A). Each pulse Capacitive sensors produce a sine (typically a 10.8-V square wave), per-
represents a discrete volume of liq- wave by generating an RF signal that mitting distances up to 3,000 ft.
uid. The number of pulses per unit is amplitude-modulated by the between the sensor and the electron-
volume is called the meter’s K-factor. movement of the rotor blades. ics without amplification.
In the inductance pick-up, the Instead of pick-up coils, Hall-effect In the water distribution industry,
permanent magnet is embedded in transistors also can be used. These mechanical-drive Woltman-type tur-
the rotor, or the blades of the rotor transistors change their state when bine flowmeters continue to be the
TRANSACTIONS Volume 4 41
Mechanical Flowmeters 3
standard. These turbine meters use a turbine meters must register whereas turbine meters intended for
gear train to convert the rotation of between 98-102% of actual rate at municipal water service are bronze or
the rotor into the rotation of a verti- maximum flow when tested. Class II cast iron. The rotor and bearing
cal shaft. The shaft passes between turbine meters must register materials are selected to match the
the metering tube and the register between 98.5-101.5% of actual rate. process fluid and the service. Rotors
section through a mechanical stuff- Both Class I and Class II meters must are often made from stainless steel,
and bearings of graphite, tungsten
carbide, ceramics, or in special cases
100
of synthetic ruby or sapphire com-
±0.15% Linearity Flow Rate
bined with tungsten carbide. In all
Nominal K Factor
98.50 cases, bearings and shafts are
Meter Coefficient K - Pulses/Gal.
99
+0.25%
designed to provide minimum fric-
42 Volume 4 TRANSACTIONS
3 Mechanical Flowmeters
Because they are very sensitive to the bility is from ±0.2% to ±0.02% over downs if accuracy is de-rated to 1%
velocity profile of the flowing stream, the linear range. of full scale (FS).
they must be profiled at several Because there are minor inconsis-
points across the flow path. tencies in the manufacturing • Sizing & Selection
Insertion turbine meters can be process, all turbine flowmeters are Turbine meters should be sized so
designed for gas applications (small, calibrated prior to shipment. The that the expected average flow is
lightweight rotor) or for liquid (larger resulting K-factor in pulses per vol- between 60% and 75% of the maxi-
rotor, water-lubricated bearings). ume unit will vary within the stated mum capacity of the meter. If the pipe
They are often used in large diame- linearity specification. It is possible, is oversized (with flow velocity under
ter pipelines where it would be cost- however, to register several K-factors 1 ft/sec), one should select a Hall-
prohibitive to install a full size meter. for different portions of the flow effect pick-up and use a meter small-
They can be hot-tapped into existing range and to electronically switch er than the line size. Flow velocities
D Bore
Flow
Alternative Flow
Dia.
Meter and Straightener
Straightening Vanes Connections
pipelines (6 in or larger) through a from one to the other as the mea- under 1 ft/sec can be insufficient,
valving system without shutting sured flow changes. Naturally, the K- while velocities in excess of 10 ft/sec
down the process. Typical accuracy factor is applicable only to the fluid can result in excessive wear. Most tur-
of an insertion turbine meter is 1% FS, for which the meter was calibrated. bine meters are designed for maxi-
and the minimum flow velocity is Barstock turbine meters typically mum velocities of 30 ft/sec.
about 0.2 ft/sec. are linear to ±0.25% AR over a 10:1 Turbine flowmeters should be
flow range. The linearity of larger sized for between 3 and 5 psid pres-
• Turbine Meter Accuracy meters is ±0.5% AR over a 10:1 flow sure drop at maximum flow. Because
Figure 3-9 shows a typical turbine- range. Turbine meters have a typical pressure drop increases with the
meter calibration curve describing nonlinearity (the turbine meter square of flow rate, reducing the
the relationship between flow and hump, shown in Figure 3-9) in the meter to the next smaller size will
K-factor (pulses/gallon). The accu- lower 25-30% of their range. Keeping raise the pressure drop considerably.
racy of turbine meters is typically the minimum flow reading above this Viscosity affects the accuracy and
given in percentage of actual rate (% region will permit linearity to within linearity of turbine meters. It is there-
AR). This particular meter has a lin- 0.15% on small and 0.25% on larger fore important to calibrate the meter
earity tolerance band of ±0.25% turbine meters. If the range of 10:1 is for the specific fluid it is intended to
over a 10:1 flow range and a ±0.15% insufficient, some turbine flow- measure. Repeatability is generally not
linearity in a 6:1 range. The repeata- meters can provide up to 100:1 turn- greatly affected by changes in viscosity,
TRANSACTIONS Volume 4 43
Mechanical Flowmeters 3
and turbine meters often are used to upstream straight-pipe runs Turbine meters also can be dam-
control the flow of viscous fluids. • 20 diameters for 90° elbow, tee, aged by solids entrained in the fluid.
Generally, turbine meters perform well filter, strainer, or thermowell; If the amount of suspended solids
if the Reynolds Number is greater than • 25 diameters for a partially open exceeds 100 mg/l of +75 micron
4,000 and less than or equal to 20,000. valve; and size, a flushing y-strainer or a
Paddlewheel
Sensor
Locknut
Flow
Pipe Tee
A) Impeller B) Paddlewheel
Because it affects viscosity, tempera- • 50 or more diameters if there are motorized cartridge filter must be
ture variation can also adversely affect two elbows in different planes or installed at least 20 diameters of
accuracy and must be compensated if the flow is spiraling or straight run upstream of the
for or controlled. The turbine meter’s corkscrewing. flowmeter.
operating temperature ranges from - In order to reduce this straight-
200 to 450°C (-328 to 840°F). run requirement, straightening vanes • New Developments
Density changes do not greatly are installed. Tube bundles or radial Dual-rotor liquid turbines increase
affect turbine meters. On low density vane elements are used as external the operating range in small line size
fluids (SG < 0.7), the minimum flow flow straighteners located at least 5 (under 2 in) applications. The two
rate is increased due to the reduced diameters upstream of the meter rotors turn in opposite directions.
torque, but the meter’s accuracy usu- (Figure 3-10). The front one acts as a conditioner,
ally is not affected. Under certain conditions, the pres- directing the flow to the back rotor.
sure drop across the turbine can cause The rotors lock hydraulically and
• Installation & Accessories flashing or cavitation. The first causes continue to turn as the flow decreases
Turbine meters are sensitive to the meter to read high, the second even to very low rates.
upstream piping geometry that can results in rotor damage. In order to The linearity of a turbine meter is
cause vortices and swirling flow. protect against this, the downstream affected by the velocity profile (often
Specifications call for 10-15 diameters pressure must be held at a value dictated by the installation), viscosity,
of straight run upstream and five equaling 1.25 times the vapor pressure and temperature. It is now possible to
diameters of straight run downstream plus twice the pressure drop. Small include complex linearization func-
of the meter. However, the presence amounts of air entrainment (100 mg/l tions in the preamplifier of a turbine
of any of the following obstructions or less) will make the meter read only flowmeter to reduce these nonlin-
upstream would necessitate that a bit high, while large quantities can earities. In addition, advances in
there be more than 15 diameters of destroy the rotor. fieldbus technology make it possible
44 Volume 4 TRANSACTIONS
3 Mechanical Flowmeters
to recalibrate turbine flowmeters Other Rotary Flowmeters the forward direction. For smaller
continuously, thereby correcting for Other types of rotary element pipes (H" to 3"), these meters are
changes in temperature and viscosity. flowmeters include propeller available only with a fixed insertion
Flow computers are capable of lin- (impeller), shunt, and paddlewheel depth, while for larger pipe sizes (4"
earization, automatic temperature designs. to 48") adjustable insertion depths
compensation, batching, calculation Propeller meters are commonly are available. The use of capacitive-
of BTU content, datalogging, and used in large diameter (over 4 in) irri- ly coupled pick-ups or Hall-effect
storage of multiple K-factors. The gation and water distribution sys- sensors extends the range of pad-
batching controller is set with the tems. Their primary trade-off is low dlewheel meters into the low-flow
desired target volume and, when its cost and low accuracy (Figure 3-11A). velocity region of 0.3 ft/sec.
totalizer has counted down to zero, AWWA Standard C-704 sets the Low-flow meters (usually smaller
it terminates the batch. Such pack- accuracy criterion for propeller than 1 in.) have a small jet orifice
ages are equipped with dribble flow, meters at 2% of reading. Propeller that projects the fluid onto a
pre-warn, or trickle-cut-off circuits. meters have a rangeability of about Pelton wheel. Varying the diameter
Whether functioning through a 4:1 and exhibit very poor perfor- and the shape of the jet orifice
relay contact or a ramp function, mance if the velocity drops below matches the required flow range
these features serve to minimize 1.5 ft/sec. Most propeller meters are and provides a flowmeter that is
splashing or overfill and to accu- equipped with mechanical registers. accurate to 1% FS and has a range-
rately terminate the batch. Mechanical wear, straightening, and ability of 100:1. Higher accuracy
conditioning requirements are the can be achieved by calibrating
• Gas Turbine & Shunt Meters same as for turbine meters. the meter and by lowering its
Gas meters compensate for the Paddlewheel flowmeters use a range. Because of the small size of
lower driving torque produced by rotor whose axis of rotation is par- the jet orifice, these meters can
the relatively low density of gases. allel to the direction of flow (Figure only be used on clean fluids and
This compensation is obtained by 3-11B). Most paddlewheel meters they incur a pressure drop of about
very large rotor hubs, very light rotor have flat-bladed rotors and are 20 psid. Materials of construction
assemblies, and larger numbers of inherently bi-directional. Several include polypropylene, PVDF, TFE
rotor blades. Gas turbine meters are manufacturers, however, use and PFA, brass, aluminum, and
available from 2" to 12" and with flow crooked rotors that only rotate in stainless steel. T
ratings up to 150,000 ft3/hr. When
operating at elevated gas pressures
(1,400 psig), a rangeability of 100:1 can References & Further Reading
be obtained in larger size meters. • OMEGA Complete Flow and Level Measurement Handbook and
Under lower pressure conditions, Encyclopedia®, OMEGA Press, 1995.
typical rangeability is 20:1 with ±1% • OMEGA Volume 29 Handbook & Encyclopedia, Purchasing Agents
linearity. The minimum upstream Edition, OMEGA Press, 1995.
straight pipe-run requirement is 20 • Flow Measurement Engineering Handbook, Miller, McGraw-Hill, 1982.
pipe diameters. • Flow Measurement, D. W. Spitzer, ISA, 1991.
Shunt flowmeters are used in gas • Flowmeters in Water Supply, Manual M33, AWWA, 1989.
and steam service. They consist of • Industrial Flow Measurement, D. W. Spitzer, ISA 1984.
an orifice in the main line and a • Instrument Engineer’s Handbook, Bela Liptak, editor, CRC Press, 1995.
rotor assembly in the bypass. These • “Turbine Flowmeter Extends Flow Range”, E. Piechota, Flow Control,
meters are available is sizes 2 in. and February, 1997.
larger and are accurate to ±2% over • Water Meters—Selection, Installation, Testing and Maintenance, Manual
a range of 10:1. M6, AWWA, 1986.
TRANSACTIONS Volume 4 45
4
V
insulating material. A pair of magnetic
coils is situated as shown in Figure 4-1,
Electrode
and a pair of electrodes penetrates ES
the pipe and its lining. If a conductive
fluid flows through a pipe of diameter Magnetic Coil
(D) through a magnetic field density Figure 4-1: The Magmeter and Its Components
46 Volume 4 TRANSACTIONS
4 Electronic Flowmeters
the magnetic coils. As a result, the zero stability problems of traditional meet these guidelines as well.
flow signal (at constant flow) will ac designs. These devices contain cir- Magnetic flowtubes and liners are
also look like a sine wave. The ampli- cuitry that periodically disrupts the available in many materials and are
tude of the wave is proportional to ac power, automatically zeroing out widely used in all the process indus-
velocity. In addition to the flow signal, the effects of process noise on the tries, including food, pharmaceutical,
noise voltages can be induced in the output signal. mining, and metals.
electrode loop. Out-of-phase noise is Today, dc excitation is used in Some liner materials (particularly
easily filtered, but in-phase noise about 85% of installations and ac Teflon®) can be damaged when pry
requires that the flow be stopped magmeters claim the other 15% when bars are used while installing it or
(with the pipe full) and the transmitter removing it from process piping.
output set to zero. The main problem They can also be damaged by over-
Varying Flux Non-Varying
with ac magmeter designs is that Flux torquing the flange bolts. Liner pro-
noise can vary with process condi- tectors are available to help prevent
tions and frequent re-zeroing is such damage.
required to maintain accuracy. Any flowtube can generally be
In dc excitation designs, a low fre- used with any transmitter offered by
quency (7-30 Hz) dc pulse is used to the same manufacturer. Depending
ac Excitation Pulsed dc
excite the magnetic coils. When the Excitation on its construction and features, the
coils are pulsed on (Figure 4-2), the cost of a 2-in. magnetic flowmeter
Figure 4-2: Excitation Methods
transmitter reads both the flow and can range from $1,500 to $5,000. This
noise signals. In between pulses, the justified by the following conditions: cost has been coming down, but is
transmitter sees only the noise signal. • When air is entrained in large still higher than that of the least
Therefore, the noise can be continu- quantities in the process stream; expensive flow sensors.
ously eliminated after each cycle. • When the process stream is a slurry Magnetic flowmeters also can be
This provides a stable zero and and the solid particle sizes are not packaged as probes and inserted into
eliminates zero drift. In addition to uniform and/or the solid phase is process pipes through taps. These
being more accurate and able to not homogeneously mixed within probes contain both the electrodes
measure lower flows, dc meters are the liquid; or and magnetic coils. The flowing
less bulky, easier to install, use less • When the flow is pulsating at a process fluid induces a voltage at the
energy, and have a lower cost of frequency under 15 Hz. electrodes, which reflects the velocity
ownership than ac meters. One new When any of the above three con- at the probe tip and not the average
dc design uses significantly more ditions exist, the output of a pulsed fluid velocity across the pipe. These
power than the earlier generations dc meter is likely to be noisy. In some magmeters are inexpensive and
and thereby creates a stronger flow- cases, one can minimize the noise retractable. Therefore, the process
tube signal. problem (hold the fluctuations with- does not have to be shut down to
Another new design uses a unique in 1% of setpoint) by filtering and install or remove them. Metering accu-
dual excitation scheme that pulses damping the output signal. If more racy is highly dependent on the rela-
the coils at 7 Hz for zero stability and than 1 to 3 seconds of damping is tionship between the measured veloc-
also at 70 Hz to obtain a stronger sig- required to eliminate the noise, it is ity and the average velocity in the pipe.
nal. Magmeter transmitters can be always better to use an ac meter.
supplied with either ac or dc power. • Electrodes
A two-wire, loop-powered dc mag- • Flowtubes, Liners, & Probes In conventional flowtubes, the elec-
netic flowmeter is also available in an The face-to-face dimensions of trodes are in contact with the process
intrinsically safe design, but its per- flanged flowtubes (lay lengths) usual- fluid. They can be removable or per-
formance is reduced because of ly meet the recommendations of the manent if produced by a droplet of
power limitations. International Organization for liquid platinum as it sinters through a
Pulsed ac meters have also been Standardization (ISO). The dimensions ceramic liner and fuses with the alu-
introduced recently, eliminating the of short-form magmeters usually minum oxide to form a perfect seal.
TRANSACTIONS Volume 4 47
Electronic Flowmeters 4
This design is preferred due to its the pipe diameter) in order to remain It is important that the conductivi-
low cost, its resistance to abrasion covered by the fluid. Compensation ty of the process fluid be uniform. If
and wear, its insensitivity to nuclear is provided for wave action and cali- two fluids are mixed and the conduc-
radiation, and its suitability for sani- bration is provided for full pipe, no tivity of one additive is significantly
tary applications because there are flow (static level), and partially filled different from that of the other
no cavities in which bacteria can pipe operation. process fluid, it is important that they
grow. On the other hand, the ceram- Another recent development is a be completely intermixed before the
ic tube cannot tolerate bending, ten- magnetic flowmeter with an unlined blend reaches the magmeter. If the
sion, or sudden cooling and cannot carbon steel flowtube. In this design, blend is not uniform, the output sig-
handle oxidizing acids or hot and the measuring electrodes mount nal will be noisy. To prevent that,
concentrated caustic. externally to the unlined flowtube pockets of varying conductivity can
30
20
)
)
Feet per second (m/s)
m
)
)
)
(2 m)
)
m
m
)
m
m
m
m
m
m
m
m
10 0 m
m
m
m
m
5m
m
m
.5
0
32 " (4
0
50
.3
0
75
2 " (15
60
10 " (2
)
0
10
(2
m
15
(6
(
5
"(
10
m
(
(
1"
2"
1
3"
1"
4"
6"
"
"
"
8"
"
"
24
16
18
12
14
00
4
9
"(
42 36
5
78 6" 60 "
6 " 48
54 "
"
" 8 72"
95 4"
"
2
90 "
1
0.01 0.1 (0.023) 1 (0.23) 10 (2.3) 100 (22.7) 1,000 (227) 10,000 (2,273) 100,000
Gallons per minute (m3/hr)
48 Volume 4 TRANSACTIONS
4 Electronic Flowmeters
lowers pumping costs and aids Installation in horizontal lines is liner wear. Liner protectors are avail-
gravity feed systems. acceptable if the pipe section is at a able to protect the leading edge of
low point and if the electrodes are the liners from the abrasive effects of
• Problem Applications not at the top of the pipe. This pre- process fluids. If the magmeter is
The magmeter cannot distinguish vents air from coming into contact installed in a horizontal pipe exceed-
entrained air from the process fluid; with the electrodes. When the ing 30 ft in length, the pipe should be
therefore, air bubbles will cause the process fluid is a slurry and the mag- supported on both sides of the meter.
magmeter to read high. If the meter is installed at a low point, it The magnetic flowmeter must be
trapped air is not homogeneously should be removed during long peri- electrically grounded to the process
dispersed, but takes the form of air ods of shutdown, so that solids will liquid. This is because the magmeter
slugs or large air bubbles (the size of not settle and coat the internals. is part of the path for any stray cur-
the electrode), this will make the If it is essential to drain the mag- rent traveling down the pipeline or
output signal noisy or even disrupt it. meter periodically, it should be pro- through the process liquid. Bonding,
Therefore, in applications where air vided with an empty tube zero by grounding the meter at both ends
entrainment is likely, the meter option. When this option is activat- to the process fluid, provides a short
should be sized so that the flow ed, the output of the transmitter will circuit for stray currents, routing
velocity under normal flow condi- be clamped to zero. Detection of them around the flowtube instead of
tions is 6-12 ft/sec. empty tube conditions is by circuitry through it. If the system is not prop-
Coating of the electrodes is anoth- connected to extra sets of elec- erly grounded, these currents can
er common magmeter problem. trodes in the flowtube. The empty create a zero shift in the magnetic
Material build-up on the inner sur- tube zero feature can also be acti- flowmeter output.
faces of the meter can electrically iso- vated by an external contact, such as Electrical bonding to the process
late the electrodes from the process
fluid. This can cause a loss of signal or High Velocity Fluid
a measurement error, either by chang-
ing the diameter of the flowtube or
by causing span and zero shifts.
Naturally, the best solution is preven-
tion. One preventive step is to size the Flow
meter such that, under normal flow
Alternative
conditions, the flowing velocity will Still Fluid Vortices
be relatively high: at least 6-12 ft/sec,
or as high as practical considering the Shear Layer
possibility of erosion and corrosion.
Another method of prevention is
to use electrodes that protrude into d
V
the flow stream to take advantage of 1D
the turbulence and washing effect. In
more severe service, a mechanical
cleaning system can be installed and
l
used intermittently or continuously
to eliminate coating and build-ups.
Figure 4-4: Vortex Meter Calculation of Flow Velocity
TRANSACTIONS Volume 4 49
Electronic Flowmeters 4
liquid. Straps are used when the pip- Theodor von Karman discovered a flagpole (which acts as a bluff
ing is electrically conductive. When that, when a non-streamlined object body); this is what causes the regular
the pipe is non-conductive or lined, (also called a bluff body) is placed in rippling one sees in a flag. Vortices
grounding rings are used. The ground- the path of a fast-flowing stream, the are also shed from bridge piers, pil-
ing ring is like an orifice plate with a fluid will alternately separate from ings, offshore drilling platform sup-
bore equal to the nominal size (inside the object on its two downstream ports, and tall buildings. The forces
diameter) of the flowtube. It is sides, and, as the boundary layer caused by the vortex-shedding phe-
installed between the flanges of the becomes detached and curls back on nomenon must be taken into
flowtube and adjacent process piping itself, the fluid forms vortices (also account when designing these struc-
on the upstream and downstream called whirlpools or eddies). He also tures. In a closed piping system, the
sides. The flowtube is bonded to the noted that the distance between the vortex effect is dissipated within a
process fluid by being connected to vortices was constant and depended few pipe diameters downstream of
the metallic grounding rings, and is solely on the size of the rock that the bluff body and causes no harm.
grounded by being wired to a good formed it.
conductor, such as a cold water pipe. On the side of the bluff body • Vortex Meter Design
In larger sizes and in exotic materi- where the vortex is being formed, A vortex flowmeter is typically
als, grounding rings can become the fluid velocity is higher and the made of 316 stainless steel or
expensive; grounding electrodes (a pressure is lower. As the vortex Hastelloy and includes a bluff body,
a vortex sensor assembly and the
transmitter electronics, although
the latter can also be mounted
remotely (Figure 4-5). They are typi-
cally available in flange sizes from
H in. to 12 in. The installed cost of
vortex meters is competitive with
Sensor
that of orifice meters in sizes under
six inches. Wafer body meters (flan-
Force on Sensor geless) have the lowest cost, while
flanged meters are preferred if the
Flow process fluid is hazardous or is at a
high temperature.
Vortex Shedder Force
Bluff body shapes (square, rectan-
Pivoting Axis gular, t-shaped, trapezoidal) and
dimensions have been experimented
with to achieve the desired charac-
Shedder Bar
teristics. Testing has shown that lin-
earity, low Reynolds number limita-
tion, and sensitivity to velocity pro-
Figure 4-5: Vortex Detecting Sensor
file distortion vary only slightly with
third electrode placed in the flowtube moves downstream, it grows in bluff body shape. In size, the bluff
for bonding with the process fluid) can strength and size, and eventually body must have a width that is a
be used instead. Another cost-saving detaches or sheds itself. This is fol- large enough fraction of the pipe
option is to use a plastic grounding ring lowed by a vortex's being formed on diameter that the entire flow partic-
with a metal electrode insert. the other side of the bluff body ipates in the shedding. Second, the
(Figure 4-4). The alternating vortices bluff body must have protruding
Vortex Flowmeters are spaced at equal distances. edges on the upstream face to fix the
As a young person fishing in the moun- The vortex-shedding phenomenon lines of flow separation, regardless of
tain streams of the Transylvanian Alps, can be observed as wind is shed from the flow rate. Third, the bluff body
50 Volume 4 TRANSACTIONS
4 Electronic Flowmeters
length in the direction of the flow to occur. The relationship between area available for flow (A):
must be a certain multiple of the vortex frequency and fluid velocity is:
bluff body width. Q = AV = (A f d B)/St
Today, the majority of vortex St = f(d/V)
meters use piezoelectric or capaci- where B is the blockage factor,
tance-type sensors to detect the Where St is the Strouhal number, f is defined as the open area left by the
pressure oscillation around the the vortex shedding frequency, d is bluff body divided by the full bore
bluff body. These detectors
respond to the pressure oscillation
with a low voltage output signal
which has the same frequency as
the oscillation. Such sensors are mod-
ular, inexpensive, easily replaced, and
can operate over a wide range of
temperature ranges—from cryo-
Downstream
genic liquids to superheated steam. Concentric
Upstream
Straight
Vortex
Meter Straight
Reducer
Sensors can be located inside the Pipe Run Pipe Run
meter body or outside. Wetted sen- Concentric
Expander
sors are stressed directly by the vor- A) Length = Size of Meter
tex pressure fluctuations and are
enclosed in hardened cases to with-
stand corrosion and erosion effects. h
External sensors, typically piezo- h>0
h
electric strain gages, sense the vortex h>0
Flow
shedding indirectly through the force Flow
Flow
exerted on the shedder bar. External
B) Upward C) Downward D) Horizontal
sensors are preferred on highly ero-
sive/corrosive applications to reduce Figure 4-6: Installation Recommendations
maintenance costs, while internal sen-
sors provide better rangeability (bet- the width of the bluff body, and V is area of the pipe. This equation, in
ter low flow sensitivity). They are also the average fluid velocity. The value turn, can be rewritten as:
less sensitive to pipe vibrations. The of the Strouhal number is deter-
electronics housing usually is rated mined experimentally, and is generally Q=fK
explosion- and weatherproof, and found to be constant over a wide
contains the electronic transmitter range of Reynolds numbers. The where K is the meter coefficient,
module, termination connections, and Strouhal number represents the ratio equal to the product (A f d B). As with
optionally a flow-rate indicator of the interval between vortex shed- turbine and other frequency-produc-
and/or totalizer. ding (l) and bluff body width (d), ing flowmeters, the K factor can be
which is about six (Figure 4-4). The defined as pulses per unit volume
• Sizing & Rangeability Strouhal number is a dimensionless (pulses per gallon, pulses per cubic
Vortex shedding frequency is directly calibration factor used to character- foot, etc.). Therefore, one can deter-
proportional to the velocity of the ize various bluff bodies. If their mine flowrate by counting the pulses
fluid in the pipe, and therefore to vol- Strouhal number is the same, then per unit time. Vortex frequencies
umetric flow rate. The shedding fre- two different bluff bodies will per- range from one to thousands of puls-
quency is independent of fluid prop- form and behave similarly. es per second, depending upon the
erties such as density, viscosity, con- Because the volumetric flowrate flow velocity, the character of the
ductivity, etc., except that the flow Q is the product of the average fluid process fluid, and the size of the
must be turbulent for vortex shedding velocity and of the cross-sectional meter. In gas service, frequencies are
TRANSACTIONS Volume 4 51
Electronic Flowmeters 4
about 10 times higher than in liquid the density or viscosity of the fluid than 10,000, error can reach 10% of
applications. differs from that of water, the meter actual flow.
The K factor is determined by the range will change. While most flowmeters continue
manufacturer, usually by water cali- In order to minimize measurement to give some indication at near zero
bration in a flow lab. Because the K noise, it is important to select a flows, the vortex meter is provided
factor is the same for liquid, gas and meter that will adequately handle with a cut-off point. Below this level,
vapor applications, the value deter- both the minimum and maximum the meter output is automatically
mined from a water calibration is valid process flows that will be measured. clamped at zero (4 mA for analog
Particles in
Upstream
Flowstream Transducer (T1)
Transmit Transducer
(Typical)
fo a
Flow
Direction a Flow
fl a Profile
Receive Transducer
(Typical) a = Refraction Angle
Downstream
A) Doppler Shift B) Transit Time Transducer (T2)
52 Volume 4 TRANSACTIONS
4 Electronic Flowmeters
to signal component or other failures. measurable. On the other hand, if liquid phase is likely to travel on the
Smart transmitters can initiate testing reduced rangeability is acceptable bottom of the pipe, and therefore
routines to identify problems with and the meter is correctly sized for the inner area of the pipe should be
both the meter and with the applica- normal flow, the vortex flowmeter kept open at the bottom. This can
tion. These on-demand tests can also can still be considered. be achieved by installing the bluff
assist in ISO 9000 verification. If the process fluid tends to coat body horizontally. Measurement
Some recently introduced vortex or build-up on the bluff body, as in inaccuracy in such applications is
flowmeters can detect mass flow. sludge and slurry service, this will about 5% of actual flow, but with
One such design measures both the eventually change the meter’s K fac- good repeatability.
vortex frequency and the vortex tor. Vortex-shedding flowmeters are The permanent pressure loss
pulse strength simultaneously. From not recommended for such applica- through a vortex meter is about half
these readings, the density of the tions. If, however, a dirty fluid has that of an orifice plate, roughly two
process fluid can be determined and only moderate amounts of non- velocity heads. (A velocity head is
the mass flow calculated to within coating solids, the application is defined as V2/g, where V is the flow
2% of span. likely to be acceptable. This was velocity and g is the gravitational
Another newer design is provided demonstrated by a 2-year test on a constant in consistent units.) If the
with multiple sensors to detect not limestone slurry. At the end of the pipe and meter are properly sized
only the vortex frequency, but also test, the K factor was found to have and of the same size, the pressure
the temperature and pressure of the changed only 0.3% from the original drop is likely to be only a few psi.
process fluid. Based on that data, it factory calibration, although the However, downsizing (installing a
determines both the density and the bluff body and flowtube were badly smaller-than-line-size meter) in order
mass flow rate. This meter offers a scarred and pitted. to increase the Reynolds can
1.25% of rate accuracy when measur- When measuring multi-phase flow increase the head loss to more than
ing the mass flow of liquids and a 2% (solid particles in gas or liquid; gas 10 psi. One should also make sure
of rate accuracy for gases and steam. bubbles in liquid; liquid droplets in that the vena contracta pressure
If knowledge of process pressure and gas), vortex meter accuracy will drop does not drop below the vapor pres-
temperature is of value for other rea-
sons, this meter provides a conve-
nient, less costly alternative to Transmitting
Receiving
Element Element
installing separate transmitters.
TRANSACTIONS Volume 4 53
Electronic Flowmeters 4
variations in process conditions and About half of all vortex meter meter. The bores of the meter, the
low wear relative to orifices or tur- installations require the “necking gaskets and the adjacent piping must
bine meters. Also, initial and mainte- down” of oversized process piping by be carefully aligned to eliminate any
nance costs are low. For these rea- concentric reducers and expanders. obstructions or steps.
sons, they have been gaining wider Even if flow straighteners are Excessive pipe vibration can be
acceptance among users. installed, some straight (relaxation) eliminated by supporting the piping
piping will still be required. on both sides of the meter, or by
• Installation Recommendations Vortex meters can be installed rotating the meter so that the sensor
When installing a vortex flowmeter in vertically, horizontally, or at any is moved out of the plane of the
an existing process where the flow angle, as long as they are kept flood- vibration. Process noise due to valve
range is not known, it is recommended ed. The meter can be kept flooded chattering, steam traps, or pumps can
Transducer B
Flanged
Flanged
End End
θ
L
V
Chordal Diametric
A) Transducer A B)
to first make some approximate by installing it in a vertical upward result in high readings or non-zero
measurements (using portable pitot flow line (Figure 4-6B). When readings under zero-flow conditions.
or clamp-on ultrasonic devices). installing the flowmeter in a down- Most meter electronics allow for
Otherwise, there is no guarantee that ward (Figure 4-6C) or horizontal increasing the noise filter settings,
a line-size vortex meter will work at all. (Figure 4-6D) flow, the downstream but increased noise reduction usually
The vortex meter requires a well- piping should be kept elevated. also decreases the low-flow sensitiv-
developed and symmetrical flow Check valves can be used to keep the ity of the meter. One option is to
velocity profile, free from any distor- piping full of liquid when there is no relocate the meter to a less noisy
tions or swirls. This necessitates the flow. Block and bypass valves are part of the process.
use of straight up- and downstream required if the replacement of the
piping to condition the flow. The sensor in the particular design Ultrasonic Flowmeters
straight length of pipe must be the requires the stopping of the flow and The speed at which sound propa-
same size as the meter (Figure 4-6) and the opening up of the process. gates in a fluid is dependent on the
its length should be about the same as Mating flanges (on the schedule fluid’s density. If the density is con-
required for an orifice installation 40 or schedule 80 mating piping) stant, however, one can use the time
with a beta ratio of 0.7 (see Chapter 2). must have the same diameter and of ultrasonic passage (or reflection)
Most vortex flowmeter manufactur- smooth bore as the flowmeter. Weld to determine the velocity of a
ers recommend a minimum of 30 pipe neck flanges are preferred, and flowing fluid.
diameters downstream of control reducing flanges should not be used. Some manufacturers produce
valves, and 3 to 4 pipe diameters The inner surface of the mating pipe transducer systems that operate in
between the meter and downstream should be free from mill scale, pits, the shear-mode, sending a single
pressure taps. Temperature elements holes, reaming scores and bumps for pulse and receiving a single pulse in
should be small and located 5 to 6 a distance of 4 diameters upstream return. Narrow-beam systems are
diameters downstream. and 2 diameters downstream of the commonly subject to walk-away (the
54 Volume 4 TRANSACTIONS
4 Electronic Flowmeters
signal completely missing the down- Because Ct /2f0cos(a) is a constant will attenuate the reflected Doppler
stream transducer). Wide-beam sys- (K), the relationship can be simpli- signal to the point where it cannot
tems overcome beam refraction and fied to: be distinguished from the back-
work better in changing liquid densi- ground noise in the pipe.
ty and temperature. With the advent V = (f0 - f1)K The reflected Doppler signal
of digital signal processing, it has is shifted from the transmitted
become possible to apply digital sig- Thus, flow velocity V (ft/sec) is frequency by approximately 6 Hz
nal coding to the transmitted signal. directly proportional to the change for every foot per second of velocity.
This can eliminate many of the prob- in frequency. The flow (Q in gpm) in a Therefore, if the flow velocity is less
lems associated with noise and varia- pipe having a certain inside diameter than 1 ft/sec, ultrasonic flowmeter-
tions in liquid chemistry. (ID in inches) can be obtained by: ing is not practical. There seems to
be no upper limit to detectable flow
• The Doppler Shift Q = 2.45V(ID)2 = 2.45[(f0 - f1)K](ID)2 velocity, as successful installations at
In 1842, Christian Doppler discovered velocities in the 40-50 ft/sec range
that the wavelength of sound per- The presence of acoustical dis- are well documented.
ceived by a stationary observer appears continuities is essential for the
shorter when the source is approach- proper operation of the Doppler • Transit Time Measurement
ing and longer when the source is mov- flowmeter. The generally accepted In this design, the time of flight of the
ing away. This shift in frequency is the rule of thumb is that for proper sig- ultrasonic signal is measured between
basis upon which all Doppler-shift nal reflection there be a minimum two transducers—one upstream and
ultrasonic flowmeters work. of 80-100 mg/l of solids with a par- one downstream (Figure 4-7B). The
Doppler flowmeter transducers ticle size of +200 mesh (+75 micron). difference in elapsed time going with
operate at 0.640 MHz (in clamp-on In the case of bubbles, 100-200 or against the flow determines the
designs) and at 1.2 MHz in wetted mg/l with diameters between +75 fluid velocity.
sensor designs. The transducer and +150 microns is desirable. If When the flow is zero, the time for
sends an ultrasonic pulse or beam either the size or the concentration the signal T1 to get to T2 is the same as
into the flowing stream. The sound of the discontinuities changes, the that required to get from T2 to T1.
waves are reflected back by such amplitude of the reflected signal When there is flow, the effect is to
acoustical discontinuities as parti- will shift, introducing errors. boost the speed of the signal in the
cles, entrained gas bubbles, or even Doppler flowmeters are often used downstream direction, while decreas-
by turbulence vortices (Figure 4-7A). to measure the flow of such fluids as ing it in the upstream direction. The
For clamp-on designs, measurement
inaccuracy ranges from ±1% to ±5%
Flow
full scale (FS).
The meter detects the velocity of
the discontinuities, rather than the
velocity of the fluid, in calculating
the flow rate. The flow velocity (V)
can be determined by:
TRANSACTIONS Volume 4 55
Electronic Flowmeters 4
where K is a calibration factor for the In the single-sensor version, the transmitted signal or attenuate the
volume and time units used, dt is the transmit and receive crystals are pot- return signal. This dramatically
time differential between upstream ted into the same sensor body, which decreases flowmeter accuracy (to
and downstream transit times, and TL is clamped onto a single point of the within only ±20%), and, in most
is the zero-flow transit time. pipe surface (Figure 4-8). In the dual- cases, clamp-on meters will not work
Theoretically, transit-time ultra- sensor version, the transmit crystal is at all if the pipe is lined.
sonic meters can be very accurate in one sensor body, while the receive Wetted transducer designs—both
Doppler and transit time are avail-
able—overcome many of these signal
1.00
attenuation limitations. The full-pipe
transit-time meter originally consisted
K = 1 Asymptote
For Flat Profile of a flanged spool section with wet-
0.95
ted transducers mounted in the pipe
wall in transducer wells opposite to
0.90
one another but at 45-degree angles
K to the flow (Figure 4-9A). Transit-time
0.85
flowmeters can be either single-path
or multiple-path designs (Figure 4-9B).
0.80
Single-path flowmeters are pro-
vided with a single pair of transduc-
0.75
ers that make a single-line velocity
K = 0.75 For Laminar Flow
measurement. They use a meter fac-
0.70
tor that is pre-determined by calibra-
104 105 106 107
1 10 100 1,000
Re tion to compensate for variations in
velocity profile and for flow section
Figure 4-11: K-Factor Variation with Reynolds Number
construction irregularities.
(inaccuracy of ±0.1% of reading is crystal is in another. In the design of multi-path
sometimes claimed). Yet the error in Clamp-on transit time meters have flowmeters, several sets of transduc-
these measurements is limited by been available since the early 1970s. ers are placed in different paths
both the ability of the signal process- Their aim is to rival the performance across the flow section, thereby
ing electronics to determine the tran- of wetted spool-piece designs, but attempting to measure the velocity
sit time and by the degree to which without the need to break the pipe or profile across the entire cross-sec-
the sonic velocity (C) is constant. The stop the process to install the meter. tion of the pipe. Multi-path instru-
speed of sound in the fluid is a func- This goal has not yet been reached. ments are used in large-diameter
tion of both density and temperature. Clamp-on Doppler flowmeters are conduits, such as utility stacks, and in
Therefore, both have to be compen- subject to interference from the pipe other applications where non-uni-
sated for. In addition, the change in wall itself, as well as from any air form flow velocity profiles exist.
sonic velocity can change the refrac- space between the sensor and the Transit-time meters can also be
tion angle (“a” in Figure 4-7B), which in wall. If the pipe wall is made of stain- used to measure both very hot (e.g.,
turn will affect the distance the signal less steel, it might conduct the trans- liquid sulfur) and very cold (liquid
has to travel. In extreme cases, the sig- mit signal far enough so that the nitrogen) fluids, and also to detect
nal might completely miss the down- returning echo will be shifted very low flows. Wetted-transducer
stream receiver. Again, this type of enough to interfere with the reading. designs for small pipes (down to H in.)
failure is known as walk-away. There are also built-in acoustic dis- are called axial or co-axial designs
continuities in concrete-lined, plas- (Figure 4-10). These devices permit
• Design Variations tic-lined, and fiberglass-reinforced transit-time measurement along a
Clamp-on ultrasonic meters come in pipes. These are significant enough path length significantly greater than
either single or dual-sensor versions. to either completely scatter the the diameter of the pipe, increasing
56 Volume 4 TRANSACTIONS
4 Electronic Flowmeters
low-flow sensitivity. the other hand, raw wastewater is not are those in HDPE pipe, because the
Originally, ultrasonic flowmeters clean enough all the time for transit- pipe wall flexes enough to change
were divided into those using the time measurement. Other waste- the diameter of the measurement
Doppler-shift principle and those water-related applications are equally area. This affects the accuracy of the
using the transit-time principle. More problematic, as the solids concentra- meter. In addition, the flexure of the
recently, flowmeters are capable of tion can be too high for either transit- pipe wall can often break the
measuring the flow of both clean flu- time or Doppler flowmeters to work acoustic coupling of the transducer
ids and of slurries with entrained properly. In still other wastewater to the outside of the pipe, causing
solids or other acoustical discontinu- applications, the problem is that the failure. Another problem area is the
ities. Microprocessors have made it acoustical absorbency of the mostly measurement of slurries that are
possible to switch automatically organic solids in wastewater attenu- acoustically absorbent, such as lime
from clean fluid mode to particulate ates the ultrasonic signals. or kaolin slurries. These applications
mode based on the "correlation fac- The use of multi-path flowmeters fail because the highly absorbent
tor". This figure of merit dramatically in raw wastewater and storm water solids attenuate the signal below
improves the accuracy of overall applications is common, while usable strength. Lower frequency
performance. In some carefully engi- Doppler or cross-correlation hybrid (0.45 MHz) sensors have been tried
neered applications, installed accu- designs are most often used to mea- for these applications, but success
racy to within 0.5% of reading has sure activated sludge and digested has been limited.
been reported. sludge flows. Multi-path, transit-time flowme-
For mining slurries, Doppler ters also measure stack gas flows in
• Applications & Performance flowmeters typically work well. power-plant scrubbers, even in very
Doppler flowmeters are not recom- Among the few problem applications large diameter stacks. T
mended for clean fluid applications.
Transit-time flowmeters, on the
other hand, are often used to mea- References & Further Reading
sure the flow of crude oils and sim- • OMEGA Complete Flow and Level Measurement Handbook and
ple fractions in the petroleum indus- Encyclopedia®, OMEGA Press, 1995.
try. They also work well with viscous • OMEGA Volume 29 Handbook & Encyclopedia, Purchasing Agents
liquids, provided that the Reynolds Edition, OMEGA Press, 1995.
number at minimum flow is either • “An Intelligent Vortex Flowmeter,” T. Kamano and others, ISA/92
less than 4,000 (laminar flow) or Proceedings, Instrument Society of America, 1992.
above 10,000 (turbulent flow). • “Application and Installation Guidelines for Volumetric and Mass
Serious non-linearities are present in Flowmeters,” D. Ginesi and C. Annarummo, ISA Transactions, 1994.
the transition region (Figure 4-11). • “Clamp-On Leak Detectors Protect Mid-Valley Line,” S. Douglas and J.
Transit-time flowmeters are the Baumoel, Pipeline & Gas Journal, April 1993.
standard for measuring cryogenic liq- • “Committee Report: Transit Time Ultrasonic Flowmeters,” AWWA
uids down to -300°C and are also Subcommittee on Ultrasonic Devices, AWWA Journal, July 1997.
used in molten metal flowmetering. • Flow Measurement Engineering Handbook, R.W. Miller, McGraw Hill, 1996.
Measurement of liquid argon, liquid • Flow Measurement, D.W. Spitzer, editor, Instrument Society of America, 1991.
nitrogen, liquid helium and molten • “Flow Sensing: The Next Generation,” D. Ginesi, Control Engineering,
sulfur have often been reported. November 1997.
Spool-section type flowmeters are • Flowmeters in Water Supply, Manual M33, AWWA, 1989.
most often used for these applica- • Industrial Flow Measurement, D.W. Spitzer, ISA, 1984
tions, especially the axial and co- • Instrument Engineers’ Handbook, Bela Liptak, editor, CRC Press, 1995.
axial designs. • Ultrasonic Clamp-On Flowmeters: Have They Finally Arrived?,” P. Espina,
Raw wastewater applications usu- Flow Control, January 1997.
ally have too few acoustic disconti- • Water Meters - Selection, Installation, Testing and Maintenance, Manual
nuities for Doppler flowmeters. On M6, AWWA, 1986.
TRANSACTIONS Volume 4 57
5
Gamma
Detector Y
Source Section YY
Magnetic
Constant
Field Coil Impeller Turbine
Field Speed Motor
own separate error to the overall Mass flow also can be measured Because a point on the equator
measurement error, and the speed of by batch weighing or by combining traces out a larger circle per day
response of such calculations is usu- an accurate level sensor with a den- than a point nearer the poles, a
ally not sufficient to detect step sitometer. Another method is to body traveling towards either pole
changes in flow. mount two d/p transmitters on the will bear eastward, because it retains
One of the early designs of self- lower part of an atmospheric tank at its higher (eastward) rotational
contained mass flowmeters operated different elevations. In this case, the speed as it passes over the more slowly
58 Volume 4 TRANSACTIONS
5 Mass Flowmeters
rotating surface of the earth. This drift centripetal acceleration directed fluid on the rotating tube can pro-
is defined as the Coriolis force. toward P and a Coriolis acceleration vide an indication of mass flowrate.
The first industrial Coriolis patents acting at right angles to ar: Naturally, rotating a tube is not prac-
date back to the 1950s, and the first tical when building a commercial
Coriolis mass flowmeters were built ar (centripetal) = w2r flowmeter, but oscillating or vibrat-
in the 1970s. These flowmeters artifi- at (Coriolis) = 2wv ing the tube can achieve the same
cially introduce a Coriolis accelera- effect. Coriolis flowmeters can mea-
tion into the flowing stream and In order to impart the Coriolis accel- sure flow through the tube in either
measure mass flow by detecting the eration (at) to the fluid particle, a the forward or reverse directions.
resulting angular momentum. force of at (dm) has to generated by In most designs, the tube is
anchored at two points and vibrated
at between these anchors.
This configuration can be envi-
sioned as vibrating a spring and mass
ar assembly. Once placed in motion, a
w
x spring and mass assembly will vibrate
at its resonant frequency, which is a
p dm v function of the mass of that assem-
bly. This resonant frequency is select-
T ed because the smallest driving force
is needed to keep the filled tube in
constant vibration.
r
TRANSACTIONS Volume 4 59
Mass Flowmeters 5
relative positions as the tubes vibrate, output from the meter. the sturdiest tubing will be thinner
causing a change in the magnetic field The natural resonance frequency than the process piping. In addition,
of the coil. Therefore, the sinusoidal of the tube structure is a function of some designs use small bore tubing,
voltage output from the coil repre- its geometry, materials of construc- which drastically increases the flow-
sents the motion of the tubes. tion, and the mass of the tube assem- ing velocity (from 5-10 ft/sec to more
When there is no flow in a two- bly (mass of the tube plus the mass than 25 ft/sec). Designs with thin
tube design (Figure 5-3A), the vibra- of the fluid inside the tube). The walls and high fluid velocities (that is,
tion caused by the coil and magnet mass of the tube is fixed. Since mass small bore tubing), may require the
drive results in identical displace- of the fluid is its density (D) multi- use of exotic materials because of
ments at the two sensing points (B1 plied by its volume (which is also erosion concerns. One will obtain the
and B2). When flow is present, fixed), the frequency of vibration can longest meter life by selecting the
Coriolis forces act to produce a sec- be related to the density of the design with the thickest wall and the
ondary twisting vibration, resulting in process fluid (D). Therefore, the slowest flow velocity that can provide
Drive (A)
No Flow No Flow
Fd v=0
Fd
Fd Fd
Flow Flow
Fi Fi v>0 Fe
Fd Fd
Fi Fi v
v
Fi Fd Fd Fi Fe
A) Fi Fi B)
a small phase difference in the rela- density of the fluid can be determined the required accuracy and range.
tive motions. This is detected at the by measuring the resonant frequency The Coriolis meter may need to be
sensing points. The deflection of the of oscillation of the tubes. (Note that made out of exotic materials because
tubes caused by the Coriolis force density can be measured at zero of corrosion considerations or to pre-
only exists when both axial fluid flow flow, as long as the tubes are filled vent pitting. Carbon or stainless steel
and tube vibration are present. with fluid and vibrating.) can often be used in process piping,
Vibration at zero flow, or flow with- Wall thickness varies considerably because a small amount of pitting
out vibration, does not produce an from design to design; however, even can be tolerated. In case of the
60 Volume 4 TRANSACTIONS
5 Mass Flowmeters
Coriolis meter, even a small amount also are available. When viscosity reference to a fixed point or plane.
of pitting cannot be tolerated information is desired, the meter pres- The tubes were excited in such a way
because the walls are thin, and pitting sure drop needs to be measured. that localized high amplitude bend-
induces stress concentrations within Other features may require informa- ing forces were created at the anchor
the tube structure. Therefore, stan- tion to be pre-programmed into the points. This resulted in severe vibra-
dard corrosion tables (based on transmitter memory. In addition, tion problems, which were alleviated
weight loss criteria) are not suitable
for selecting Coriolis tube materials,
and the stricter guidelines of the
manufacturers must be used.
FLOW VARIABLES
CONFIGURE METER
• Mass Flow Rate
F1 F2 F3 F4 • Mass, Volume,
• Transmitter Designs • Mass Flow Total
• Volumetric Flow Rate
Timebase Units
• Meter Constants
Transmitters can operate on either ac • Volumetric Flow Total
• Configure Outputs
• Density
+ 1 2 3 • Scale Outputs
or dc power and require separate • Temperature
- 4 5 6 • Configure Inputs
• % Solids
wiring for the power supply and for • Dry Solids Mass
7
.
8 9
0
their output signals. The Coriolis Flow Rate
• Dry Solids Total DIAGNOSTICS &
flowmeter transmitter can be integrally SIMULATION
or remotely mounted (Figure 5-4). The CONFIGURABLE
transmitter controls the operation of INPUTS/OUTPUTS
KEYBOARD SECURITY
the driver and processes and trans- ANALOG & FREQUENCY
mits the sensor signals. The calibra- OUTPUTS
TRANSACTIONS Volume 4 61
Mass Flowmeters 5
velocities (up to 50 ft/sec) still result- the number of suppliers and con- tube temperature is continuously
ed in premature meter failure, includ- tributed to the development of a measured by an RTD element and is
ing potentially catastrophic spills new generation of Coriolis meters used to continuously compensate
when the meter was used on corro- that are as reliable and rugged as tra- for variations in tube elasticity.
sive and erosive services. In addition, ditional volumetric flowmeters. The Coriolis mass flowmeters usually
the unrecovered head losses were new designs operate at lower fluid are calibrated on water, because the
high (sometimes over 50 psid), and velocities (below 10 ft/sec) and at constants are valid for all other liq-
accuracy was not high enough to lower pressure drops (under 12 psid), uids. Calibration for density is usually
allow users to convert batch processes can be installed in any orientation, done by filling the tubes with two or
into continuous ones. and provide longer service life on more (stagnant) calibration fluids of
More recent design improve- slurry, viscous, corrosive, or erosive known densities.
ments include the introduction of a services. The tubes are vibrated well
variety of new tube shapes, includ- below their endurance limits, and • Accuracy & Rangeability
ing ones that do not split the flow typically are made of stainless steel, Coriolis meters provide 0.1-2% of
(Figure 5-3B) and the use of multiple Hastelloy, and titanium. rate inaccuracy over a mass flow
drivers (Figure 5-5A). Thick-walled range of up to 100:1. In general,
tubing (five times thicker than early • Interferences curved tube designs provide wider
designs), the use of full bore diame- The effect of the Coriolis force on rangeability (100:1 to 200:1), while
ters and heavy manifolds to isolate the vibrating tube is small. Full-scale straight-tube meters are limited to
the tube structure from stresses flow might cause a deflection of 30:1 to 50:1 and their accuracy is
induced from piping connections, only 0.001 inch. To obtain a flow lower. Overall meter error is the sum
and flowtube housings that double rangeability of 100:1, sensors must of base inaccuracy and zero-shift
as secondary containment vessels be able to detect deflections to an error, the error attributable to the
have all contributed to improved accuracy of 0.000001 inch in indus- irregular output signal generated at
performance. trial environments where the zero flow conditions. Zero-shift
In some designs, torsional stresses process pressure, temperature, and error becomes the dominant portion
replaced bending, in order to prevent fluid density are all changing, and of total error at the lower end of the
the concentration of stresses that where pipe vibration interferes with flow range, where the error is
can lead to tube cracking (Figure 5- measurement. between 1% and 2% of rate. Some
5B). In other designs, the effects of The elasticity of metal tubes manufacturers state the overall
pipeline vibration have been mini- changes with temperature; they accuracy as a percentage of rate for
mized by mounting the tube struc- become more elastic as they get the upper portion of the flow range
tures transverse to the pipeline. warmer. To eliminate the corre- and as a percentage of span for the
These improvements increased sponding measurement error, the lower portion, while others state it
62 Volume 4 TRANSACTIONS
5 Mass Flowmeters
as a percentage of rate plus a zero- ambient temperature and pressure Downsizing (using a meter that is
shift error. There is a fair amount of fluctuations alter these forces, perfor- smaller than the pipe) is acceptable
“specmanship,” and one must read mance may be affected and re-zeroing when the pipe is oversized and the
sales literature carefully when com- of the meter may be required. process fluid is clean with a low vis-
paring different devices. Variations in the density of the cosity. On corrosive, viscous, or abra-
When used for density measure- process fluid can affect the frequency sive slurry services, downsizing is not
ment, the typical error range of a transfer function of mechanical sys- recommended. A list of acceptable
Coriolis measurement is 0.002- tems, necessitating the re-zeroing of flow tube sizes and corresponding
0.0005 g/cc. older designs to protect them from pressure drops, inaccuracies, and flow
Errors are caused by air or gas degraded performance. Because of velocities can be obtained from soft-
pockets in the process fluid. In the their tube configurations, newer ware provided by the manufacturer.
case of homogeneously dispersed designs are unaffected by density Different Coriolis meters incur dif-
small bubbles, more power is changes over wide ranges of specific ferent pressure drops, but in general
required to vibrate the tubes, where- gravity variations. they require more than traditional
as, if the gas phase separates from volumetric meters, which usually
the liquid, a damping effect on tube • Sizing & Pressure Drop operate at less than 10 psid. (The
vibration (and, consequently, error) Because of the wide rangeability of yearly electricity cost of pumping
develops. Small voids also cause Coriolis flowmeters (30:1 to as high as 1 gpm across a differential of 10 psid
noise because of the sloshing of the 200:1), the same flow can be mea- is about $5 U.S.). This higher head loss
process liquid in the tubes. Larger sured by two or three different sized is due to the reduced tubing diameter
Support
Mass Flowtube
(Typical)
Enclosure
Mass Tube Enclosure
Flanges Flow
Support
Pipe/Flowtube Junction Direction
Arrow
NOTE:
Distance Between
Pipe/Flowtube
Junction and
Support
NOTE: Distance Between
Flow
Must Not
Pipe/Flowtube Junction and
Direction Arrow Exceed 15 Inches
Support Must Not
Exceed 15 Inches
A) Horizontal B) Vertical
'U' Rest 'V' Rest 'V' Bolt
Inverted Pipe
'V' Block Clamp
C) Pipe Supports Clamp Hanger Clamp (Can Be Inverted)
voids will raise the energy required flow tubes. By using the smallest and the circuitous path of flow.
to vibrate the tubes to excessive lev- possible meter, one will lower the Besides pumping costs, head loss can
els and may cause complete failure. initial cost and reduce coating build- be of concern if the meter is installed
Because the flowtube is subjected up, but will increase erosion/corro- in a low-pressure system, or if there is
to axial, bending, and torsional forces sion rates and head loss, increasing a potential for cavitation or flashing,
during meter operation, if process or pumping and operating costs. or if the fluid viscosity is very high.
TRANSACTIONS Volume 4 63
Mass Flowmeters 5
The viscosity of non-Newtonian intrinsically safe circuits between the still remaining homogeneously dis-
fluids is a function of their flowing flow tube and the transmitter. persed. Gas content in low viscosity
velocity. Dilettante fluids, for exam- Therefore, the amount of driving fluids, like milk, will separate at con-
ple, increase their apparent viscosity power that can be delivered to the centrations as low as 1%.
(resistance to flow) as their velocity flow tube is limited. The cost of an average-sized
is increased. This apparent viscosity When fluid is unloaded from tank (under 2 in.) Coriolis flowmeter is
can be drastically higher than their trucks, drums, or railroad cars, slug between $4,000 and $5,000. These
Air
Eliminator
Check
Valve
Pump Meter Load
Tank Bidirectional
Valve Back Pressure
Flow Valve
Unload
C) Typical Installation
Valve Check Valve
viscosity when stagnant. In order to flow can occur, making the meter out- mass flowmeters provide short pay-
provide suppliers with data on the put unpredictable. If a slug-flow recov- back periods on applications where
flowing viscosity in a particular pipe, ery feature is provided in the transmit- measurement accuracy lowers pro-
head loss per foot of pipe (used in ter, it will stop the measurement when duction costs (bathing, billing) or
pump sizing calculations) can be slug flow is detected by the excessive where multiple measurements
used as an approximation. drive power drawn or by the drop in (including density, temperature, pres-
process density (reduction in sensor sure) are needed. On the other hand,
• Applications & Limitations output amplitude). they may not be competitive when
Coriolis mass flowmeters can detect The amount of air in the process used in simple flow measurement
the flow of all liquids, including fluid that can be tolerated by a meter applications where volumetric sen-
Newtonian and non-Newtonian, as varies with the viscosity of the fluid. sors are sufficient and where
well as that of moderately dense Liquids with viscosities as high as repeatability is more important than
gases. Self-draining designs are avail- 300,000 centipoise can be metered precision. The ability to extract data
able for sanitary applications that with Coriolis meters. Gas content in on total mass charged, solids rate,
meet clean-in-place requirements. such highly viscous liquids can be as percent solids, and viscosity from a
Most meters are provided with high as 20% with the small bubbles single instrument does lower the
64 Volume 4 TRANSACTIONS
5 Mass Flowmeters
total cost of measurement, improves cleaned by mechanical means, while of tube rupture, particularly if the
process control, and provides redun- curved-tube designs are usually process fluid is likely to vaporize
dancy for other instruments. washed out using cleaning solutions under such conditions. If that is the
Continuous tube designs are gen- at velocities in excess of 10 ft/sec. case, secondary containment hous-
erally preferred for slurry and other Straight-tube designs also are pre- ings can be ordered that enclose the
multi-phase fluid applications. The ferred for sanitary applications due entire flow tube, including its hous-
total flow is divided by splitters in to self-draining requirements. ing. Such secondary containment
split-tube designs, and the resulting Long, bent tubes twist more easily enclosures can be provided with rup-
two streams do not have to be at than do short, straight tubes and ture disks or pressure relief valves,
exactly the same mass flow rate to therefore will generate stronger sig- and with drains or vents.
maintain accuracy (they do, however, nals under the same conditions. In
need to have the same density). general, curved-tube designs provide • Installation Recommendations
Different densities in the two paral- wider rangeability (100:1 to 200:1), There are no Reynolds number limi-
lel tubes imbalance the system and while straight-tube meters are limited tations associated with Coriolis
create measurement errors. to 30:1 to 50:1, with lower accuracy. meters. They are also insensitive to
Therefore, if there is a secondary Straight-tube meters are more velocity profile distortion and swirl.
phase in the stream, a simple flow immune to pipeline stresses and Therefore, there is no requirement
splitter may not evenly distribute vibration, are easy to install, require for straight runs of relaxation piping
the flow between the two tubes. less pressure drop, can be mechani- upstream or downstream of the
Continuous tube designs are also cally cleaned, are more compact, and meter to condition the flow.
preferred for measuring fluids that require less room for installation. The meter should be installed so
can coat and/or clog the meter. They are also preferred on services that it will remain full of liquid and
Temperature
Difference
Indicator
Flow
Mass Flow
Rate Film
Constant
Power
Source (q)
Upstream
Resistance
Resistance
Resistance
Temperature
Heater
Thermometer Heater Thermometer Sensor
Supplies
Downstream
Measures Tf q Temperature Sensor
T1 T2 Measures Tw
Continuous tubes, if sized to pass the where the process fluid can solidify so air cannot get trapped inside the
largest solid particles in the process at ambient temperatures. tubes. In sanitary installations, the
fluid, are less likely to clog and are Not all meter housings are meter must also drain completely. The
easier to clean. designed to withstand and contain most desirable installation is in vertical
Straight-tube designs can be the pressurized process fluid in case upward flow pipes (Figure 5-6B), but
TRANSACTIONS Volume 4 65
Mass Flowmeters 5
installations in horizontal lines (Figure If air bubbles are likely to be present different materials at different
5-6A) are also acceptable. Installations in the process fluid, it is recommended times; and
where the flow is downward in a verti- to install an air release upstream of the • Allowing foam formation by high
cal pipe are not recommended. meter. System design characteristics turbulence in high velocity fluids.
In newer Coriolis designs, normal that can result in the presence of air It is recommended to install
pipe vibration should not affect the (and which can often be eliminated at (upstream of the meter) strainers, fil-
performance of the Coriolis meter if the design stage) include: ters or air/vapor eliminators as
it is properly supported by the • Common piping used for pumping required to remove all undesirable sec-
process piping (Figure 5-6C). No spe- into and out of storage tanks; ondary phases. Figure 5-7C illustrates
cial supports or pads are needed for • Allowing the formation of a vortex an air eliminator installation. Its func-
the flow tube, but standard piping in stirred vessels under low-level tion is to slow the velocity of the liq-
supports should be located on either conditions; uid, thereby allowing time for the
side of the meter. If the installation • Allowing air leakage through pack- entrained air to separate and be
instructions require special hardware ing glands of pumps that develop removed by venting. The rise and fall
or supports, the particular meter high vacuums on the suction side of the liquid level in the eliminator
design is likely to be sensitive to (this can occur when pumping due to the accumulation of free air
vibration, and the pulsation dampen- from underground storage); closes and opens the vent valve and
ers, flexible connectors, and mount- • Vaporization of stagnant process discharges the air (Figure 5-7A&B).
ing/clamping attachments recom- fluid in pipes exposed to the sun; Prior to zeroing the meter, all air
mended by the manufacturer should • High valve pressure drops causing should be removed. This can be
be carefully installed. vaporization and flashing; accomplished by circulating the
Meter
Zero
Flow
Temperature of Tube
Small
Bypass
Flow
Orifice
TC-1 TC-2
Main
Orifice
L/2 0 L/2
Length of Tube
If your application requires that • Allowing the pipe to drain for any process fluid through the meter for
you install two Coriolis flowmeters reason, including lack of check several minutes at a velocity of
in series or mount two Coriolis meters valves; approximately 2-6 ft/sec. On batching
near each other, the manufacturer • Allowing storage tanks, trucks, or or other intermittent flow applica-
should be consulted to prevent railroad cars to drain completely; tions, the meter should stay flooded
crosstalk between the two units. • Using the same pipe for pumping so that it does not need to be
66 Volume 4 TRANSACTIONS
5 Mass Flowmeters
repurged. All meters should be so standard of higher accuracy, such as must be installed by the manufacturer.
installed so they can be zeroed while a dead-weight calibrated weigh tank. When flowmetering is not required,
filled with liquid. Before Coriolis meters, the reference the Coriolis meter can be used solely
When zeroing the meter, any standard was expected to be an as a densitometer. In that case, to min-
associated pumps or other equip- order of magnitude more accurate imize cost, usually a small (H in.) meter
ment should be running so that their than the meter being calibrated; is installed in a by-pass line. Such a
Flow
Air Velocity Sensor
Temperature Sensor
0.032" Dia.
0.125" Dia.
0.25" Dia.
.125"
.125" Inlet
Flow Element
Temperature
.938 Screens (Heated) Compensator
noise can be zeroed out. This can be however, due to the high accuracy of configuration is acceptable only in
achieved in most cases by locating a Coriolis meters, this is rare. clean services that will not clog the
shut-off value downstream of the In less critical installations (where small bore of the meter. In addition, a
meter and either operating the weigh tanks are not used), volumetric restriction must be placed in the main
pump with its discharge blocked, provers or master meters (typically piping (between the by-pass taps) to
which is acceptable with centrifugal another Coriolis or a turbine meter ensure a flow through the meter.
pumps for a short period, or by calibrated at a flow laboratory) are
opening the pump bypass on posi- used. When a volumetric reference is Thermal Mass Flowmeters
tive displacement pumps. Valves used in calibrating a mass flowmeter, Thermal mass flowmeters also mea-
used in zeroing the meter should the fluid density must be very pre- sure the mass flowrate of gases and
provide tight shut-off; double-seat- cisely determined. liquids directly. Volumetric measure-
ed valves are preferred. Control valves should be installed ments are affected by all ambient
Meters that are expected to be downstream of the meter to increase and process conditions that influ-
calibrated in-line must be provided the back-pressure on the meter and ence unit volume or indirectly affect
with block and bypass valves so that lower the probability of cavitation or pressure drop, while mass flow mea-
the reference standard (master) flashing. surement is unaffected by changes in
meter can be installed and discon- When the process fluid must be viscosity, density, temperature, or
nected without interrupting the held at higher temperatures, some pressure.
process. The requirements for in-line Coriolis meters can be supplied with Thermal mass flowmeters are
calibration (for ISO 9000 verifica- steam jackets. As an alternative, elec- often used in monitoring or control-
tion) consist of comparing the out- trical heating tape can be added to ling mass-related processes such as
put of the meter against a reference the housing. Jackets or heating tapes chemical reactions that depend on
TRANSACTIONS Volume 4 67
Mass Flowmeters 5
68 Volume 4 TRANSACTIONS
5 Mass Flowmeters
measuring the wall temperature at of the process fluid must stay con- the speed of response of the mea-
this location further away from the stant when using this design. surement. On the other hand,
heater. This heat transfer process is because of the small size, filters are
non-linear, and the corresponding • Bypass-Type Design necessary to prevent plugging. One
equation differs from the one above The bypass version of the thermal serious limitation is the high pressure
as follows: mass flowmeter was developed to drop (up to 45 psi) needed to develop
measure larger flow rates. It consists laminar flow. This is typically accept-
m0.8 = Kq/(Cp (Tw - Tf)) of a thin-walled capillary tube able only for high pressure gas appli-
(approximately 0.125 in diameter) and cations where the pressure needs to
This flowmeter has two operating two externally wound self-heating be reduced in any case.
modes: one measures the mass flow resistance temperature detectors This is a low accuracy (2% full
by keeping the electric power input (RTDs) that both heat the tube and scale), low maintenance, and low
constant and detecting the tempera- measure the resulting temperature cost flowmeter. Electronic packages
ture rise. The other mode holds the rise (Figure 5-9A). The meter is placed within the units allow for data acqui-
temperature difference constant and in a bypass around a restriction in the sition, chart recording, and computer
measures the amount of electricity main pipe and is sized to operate in interfacing. These devices are popular
6 6
5 5
60°
4 4 60°
1 2 3 4 5 6
3 3
2 2
1 1
A) B)
needed to maintain it. This second the laminar flow region over its full in the semiconductor processing
mode of operation provides for a operating range. industry. Modern day units are also
much higher meter rangeability. When there is no flow, the heaters available as complete control loops,
Heated-tube designs are generally raise the bypass-tube temperature to including a controller and automatic
used for the measurement of clean approximately 160°F above ambient control valve.
(e.g., bottled gases) and homoge- temperature. Under this condition, a
neous (no mixtures) flows at moder- symmetrical temperature distribu- • Air Velocity Probes
ate temperature ranges. They are not tion exists along the length of the Probe-style mass flowmeters are
recommended for applications tube (Figure 5-9B). When flow is tak- used to measure air flows and are
where either the fluid composition ing place, the gas molecules carry the insensitive to the presence of mod-
or its moisture content is variable, heat downstream and the tempera- erate amounts of dust. They maintain
because the specific heat (Cp) would ture profile is shifted in the direction a temperature differential between
change. They are not affected by of the flow. A Wheatstone bridge two RTDs mounted on the sensor
changes in pressure or temperature. connected to the sensor terminals tube. The upper sensor measures the
Advantages include wide rangeability converts the electrical signal into a ambient temperature of the gas
(the ability to measure very low mass flow rate proportional to the (Figure 5-10A) and continuously
flows) and ease of maintenance. The change in temperature. maintains the second RTD (near the
temperature difference (or heater The small size of the bypass tube tip of the probe) at 60°F above ambi-
power), flowmeter geometry, thermal makes it possible to minimize electric ent. The higher the gas velocity, the
capacity, specific heat, and viscosity power consumption and to increase more current is required to maintain
TRANSACTIONS Volume 4 69
Mass Flowmeters 5
the temperature differential. chemical and petrochemical indus- rate of cooling corresponds to the
Another version of the velocity tries, research and development mass flowrate.
probe is the venturi-type thermal applications, gas chromatography, The circuitry of the heated sensing
mass flowmeter, which places a and filter and leak testing. While hot- element is controlled by one of two
heated mass flow sensor at the mini- wire anemometers are best suited for types of solid-state electronic circuits:
mum diameter of a venturi flow ele- clean gases at low velocities, venturi constant-temperature or constant-
ment and a temperature compensa- meters can also be considered for power. The constant-temperature sen-
tion probe downstream (Figure 5- some liquid (including slurry) flow sor maintains a constant temperature
10B). An inlet screen mixes the flow
to make the temperature uniform.
This design is used for both gas and
liquid measurement (including slur-
ries), with flow range a function of
the size of the venturi. Pressure drop
is relatively low and precision is
dependent upon finding the proper
probe insertion depth.
A flow switch version is also avail-
able that contains two temperature
sensors in the tip. One of the sensors
is heated and the temperature differ-
ence is a measure of velocity. The Air velocity probe provides 1.5% accuracy for local flow rate measurement.
switch can be used to detect high or
low flow within 5%. applications. Thermal mass flowme- differential between a heated sensor
ters are well suited for high range- and a reference sensor; the amount of
• Uses & Limitations ability measurements of very low power required to maintain the differ-
Thermal mass flowmeters can have flows, but also can be used in mea- ential is measured as an indication of
very high rangeability and reasonable suring large flows such as combus- the mass flow rate.
accuracy, but they also have serious tion air, natural gas, or the distribu- Constant-temperature anemome-
limitations. Potential problems include tion of compressed air. ters are popular because of their
the condensation of moisture (in satu- high-frequency response, low elec-
rated gases) on the temperature Hot-Wire Anemometers tronic noise level, immunity from
detector. Such condensation will The term anemometer was derived sensor burnout when airflow sud-
cause the thermometer to read low from the Greek words anemos, denly drops, compatibility with hot-
and can lead to corrosion. Coating or “wind,” and metron, “measure.” film sensors, and their applicability
material build-up on the sensor also Mechanical anemometers were first to liquid or gas flows.
will inhibit heat transfer and cause developed back in the 15th century Constant-power anemometers do
the meter to read low. Additional to measure wind speed. not have a feedback system.
potential sources of error include A hot-wire anemometer consists Temperature is simply proportional
variations in the specific heat caused of an electrically heated, fine-wire to flowrate. They are less popular
by changes in the gas’s composition. element (0.00016 inch in diameter because their zero-flow reading is
Some common gas-flow applica- and 0.05 inch long) supported by not stable, temperature and velocity
tions for thermal mass flowmeters needles at its ends (Figure 5-11). response is slow, and temperature
include combustion air measurement Tungsten is used as the wire materi- compensation is limited.
in large boilers, semiconductor al because of its strength and high
process gas measurement, air sam- temperature coefficient of resis- • Air Duct Traversing
pling in nuclear power plants, tance. When placed in a moving Anemometers are widely used for air
process gas measurements in the stream of gas, the wire cools; the duct balancing. This is accomplished
70 Volume 4 TRANSACTIONS
5 Mass Flowmeters
TRANSACTIONS Volume 4 71
6
simply the determination of the one will substantially shorten the list PT
presence or absence of a fluid at a of sensors to consider. The selection
particular elevation. is further narrowed by considering
only those designs that can be pro- Mass ± 0.2%
Density ± 0.3%
Level Sensor Selection vided in the required materials of Volume ± 0.28%
When determining what type of construction and can function at the
Intelligent Multi-Transmitter Package
Figure 6-2:
level sensor should be used for a required accuracy, operating temper-
given application, there are a series ature, etc. (Table 4). When the level which the level can change (due to
of questions that must be answered: to be measured is a solid, slurry, filling or discharging) and disregard-
• Can the level sensor be inserted foam, or the interface between two ing any change that occurs faster
72 Volume 4 TRANSACTIONS
6 A Level Measurement Orientation
SLURRY/SLUDGE
MAX. TEMP. (°F)
NONCONTACT
AVAILABLE AS
INACCURACY
INTERFACE
POWDER
VISCOUS
CHUNKY
STICKY
CLEAN
FOAM
TYPE
LIMITATIONS
Air
Bubblers
UL
1-2% FS
G
F
P
F
Introduces foreign substance into process;
high maintenance
Capacitance
2,000
1-2% FS
G
F-G
F
G-L
P
F
F
P
Interface between conductive layers
and detection of foam is a problem
Conductivity Switch
1,800
1/ in
F
P
F
L
L
L
L
L
Can detect interface only between
8
conductive and nonconductive liquids.
Field effect design for solids
Diaphragm
350
0.5% FS
G
F
F
F
F
P
Switches only for solid service
Only extended diaphragm seals or
Differential Pressure
1,200
0.1% AS
E
G-E
G
P
repeaters can eliminate plugging.
Purging and sealing legs are also used
Displacer
850
0.5% FS
E
P
P
F-G
Not recommended for sludge or
slurry service
Moving parts limit most designs to
Float
500
1% FS
G
P
P
F
clean service. Only preset density
floats can follow interfaces
Laser
UL
X
0.5 in
L
G
G
F
F
F
F
Limited to cloudy liquids or bright solids in
tanks with transparent vapor spaces
Level Gages
700
0.25 in
G
F
P
F
Glass is not allowed in some processes
Microwave Switches
400
X
0.5 in
G
G
F
G
G
G
F
Thick coating is a limitation
Optical Switches
260
X
0.25 in
G
F
E
F-G
F
F
P
F
Refraction-type for clean liquids only;
reflection-type requires clean vapor space
Radar
450
0.12 in
G
G
F
P
F
P
Interference from coating, agitator
X
P
blades, spray, or excessive turbulence
Radiation
UL
X
0.25 in
G
E
E
G
F
G
E
E
Requires NRC license
Resistance Tape
225
0.5 in
G
G
G
Limited to liquids under near-atmospheric
pressure and temperature conditions
Rotating Paddle Switch
500
1 in
G
F
P
Limited to detection of dry, non-corrosive,
low-pressure solids
Slip Tubes
200
0.5 in
F
P
P
An unsafe manual device
Tape-Type Level Sensors
300
0.1 in
E
F
P
G
G
F
F
Only the inductively coupled float is suited
for interface measurement. Float hangup is
a potential problem with most designs
Thermal
850
0.5 in
G
F
F
P
F
Foam and interface detectiom is limited by
the thermal conductives involved
TDR/PDS
221
3 in
F
F
F
G
G
F
Limited performance on sticky
process materials
Ultrasonic
300
X
1% FS
F-G
G
G
F-G
F
F
F
G
Presence of dust, foam, dew in vapor
space; sloping or fluffy process material
interferes with performance
Vibrating Switches
300
0.2 in
F
G
G
F
F
G
G
Excessive material buildup can prevent
operation
than that. multi-transmitter systems when it is Figure 6-2 is capable of simultane-
The relationship between level desirable to: ously measuring volume (level), mass
and tank volume is a function of the • Detect the true level, while either (weight), and density, all with an
cross-sectional shape of the tank. the process temperature or densi- accuracy of 0.3% of full span.
With vertical tanks, this relationship ty varies;
is linear, while with horizontal or • Measure both level and density; Boiling & Cryogenic Fluids
spherical vessels, it is a non-linear and When a d/p cell is used to measure
relationship (Figure 6-1). • Measure the volume and the mass the level in a steam drum, a reverse-
If the level in a tank is to be (weight) in the tank. acting transmitter is usually installed
inferred using hydrostatic pressure By measuring one temperature and (Figure 6-3). An uninsulated condens-
measurement, it is necessary to use three pressures, the system shown in ing chamber is used to connect the
TRANSACTIONS Volume 4 73
A Level Measurement Orientation 6
Displacer
1
2
2
2
-
-
3
2
-
-
-
-
-
-
-
-
-
-
Float
1
-
2
-
-
-
3
-
-
-
-
-
-
-
-
-
-
-
Float/Tape
3
1
-
-
-
-
-
3
-
-
-
-
-
-
-
-
-
-
Paddlewheel
-
-
-
-
-
-
3
-
-
-
2
-
1
-
3
-
2
-
Weight/Cable
3
1
-
-
-
-
-
1
-
1
-
1
-
1
-
1
-
1
Glass
1
1
2
2
3
3
3
3
-
-
-
-
-
-
-
-
-
-
Gages
Magnetic
1
1
-
-
3
3
3
3
-
-
-
-
-
-
-
-
-
-
Inductive
-
-
-
-
-
-
2
-
-
-
2
2
2
2
2
2
3
3
Microwave
1
1
-
-
-
-
1
1
-
-
1
2
1
1
1
1
1
1
Radiation
1
1
-
-
-
-
1
1
-
-
1
1
1
1
1
1
1
1
Sonar
-
-
2
2
-
-
-
3
1
1
-
-
-
-
-
-
-
-
Sonic Echo
Sonic
1
1
3
3
-
-
1
1
2
2
-
3
1
1
1
1
2
1
Ultrasonic
1
2
2
2
-
-
1
2
1
1
-
3
2
2
1
2
2
2
Thermal
1
-
1
-
2
-
2
-
-
-
-
-
-
-
-
-
-
-
Vibration
2
-
3
-
-
-
2
-
1
-
1
-
1
-
2
-
1
-
74 Volume 4 TRANSACTIONS
6 A Level Measurement Orientation
minimize the turbulence caused by episode, an automatic washing spray period of time, one aims to keep the
the simultaneous boiling and re- is activated. foam level below a maximum. In
condensing occurring at the liquid- When the sludge or slurry level is other processes, it is desirable to sep-
vapor interface. detected continuously, one of the arately control both the liquid level
goals is to eliminate dead-ended cavi- beneath the foam and the thickness
Sludge, Foam, & Molten Metals ties where the sludge might settle. In of the foam. In the paper industry,
Many process fluids are aggressive or addition, all surfaces which are beta radiation detectors are used for
difficult to handle and it’s best to exposed to the process fluid should such applications (Kraft processing),
while other industries detect the
degree of foaming indirectly (by mea-
1:1 Repeater suring related variables, such as heat
input or vapor flow), or they use
Pv
capacitance, conductivity, tuning
fork, optical, or thermal switches, all
provided with automatic washers.
To
Measuring the level of molten
Controller
glass or metals is another special
application. The most expensive
(but also most accurate) technique
available is proximity capacitance-
based level measurement, which
Differential
Pressure
can provide a resolution of 0.1 mm
Transmitter over a range of 6 in. Laser-based
A) Level Switch B) Continuous Detection systems can provide even better
resolution from distances up to 2 ft.
Figure 6-5: Level-Instrumentation Design For Sludges, Slurries, & High Viscosity If such high resolution is not
required and cost is a concern, one
avoid physical contact with them. be covered with Teflon®. Figure 6-5B can make a float out of refractory
This can be accomplished by placing shows such an installation, employing material and attach a linear variable
the level sensor outside the tank Teflon®-coated extended diaphragms differential transformer (LVDT), or
(weighing, radiation) or locating the to minimize material buildup. make a bubbler tube out of refrac-
sensor in the vapor space (ultrasonic, In strippers, where the goal is to tory material and bubble argon or
radar, microwave) above the process drive off the solvent in the shortest nitrogen through it. T
fluid. When these options are not
available or acceptable, one must aim
to minimize maintenance and physi- References & Further Reading
cal contact with the process fluid. • OMEGA Complete Flow and Level Measurement Handbook and
When the process fluid is a sludge, Encyclopedia®, OMEGA Press, 1995.
slurry, or a highly viscous polymer, • OMEGA Volume 29 Handbook & Encyclopedia, Purchasing Agents
and the goal is to detect the level at Edition, OMEGA Press, 1995.
one point, the design shown in Figure • Instrument Engineer’s Handbook, Bela G. Liptak, editor, CRC Press, 1995.
6-5A is commonly considered. The • Instrumentation for Process Measurement and Control, Third Edition, N.
ultrasonic or optical signal source A. Anderson, Chilton, 1980.
and receiver typically are separated • Measurement and Control of Liquid Level, C. H. Cho, Instrument Society
by more than six inches so that the of America, 1982.
process fluid drains freely from the • Principles of Industrial Measurement for Control Applications, E. Smith,
intervening space. After a high-level Instrument Society of America, 1984.
TRANSACTIONS Volume 4 75
7
Dry
Leg
dp = h (SG)
h
LI LI
LI LT LT
dp = h (SG)
Pneumatic
Pneumatic
A) Local B) Remote Supply C) Compensation Supply
76 Volume 4 TRANSACTIONS
7 Pressure/Density Level Instrumentation
Low
Zero
Pressure
Side Zero Process
Tank
High
Pressure
Side
Repeated
Repeated
Positive
Vacuum
Pressure Pressure
Any changes in leg fill level (due to tank. The required piping and valv- cell. If the tank connection is subject
leakage or vaporization) introduce ing must always be provided on to material build-up or plugging,
error into the level measurement. If both the tank and the reference leg extended diaphragm Type 1:1 repeaters
the specific gravity of the filling fluid side of the d/p cell, so that draining can be considered for the service
for the wet leg is greater than that of and flushing operations can easily (Figure 7-2).
the process fluid, the high pressure be performed. When a wet refer- While repeaters eliminate the
side should be connected to the ref- ence leg is used, a low thermal errors caused by wet legs, they do
erence leg and the low to the tank. expansion filling fluid should be introduce their own errors as a func-
If the condensate can be used to selected. Otherwise, the designer tion of the pressure repeated. For
fill the reference leg, a condensate must correct for the density varia- example, at 40 psig, repeater error is
pot can be mounted and piped both tions in the reference leg caused by about 2 in. At 400 psig, it is 20 in. In
to the high level connection of the ambient temperature variations. many applications, the former is
tank and to the top of the vapor If smart transmitters are used and acceptable but the latter is not.
space. The condensate pot must be if the filling fluid data is known, wet-
mounted slightly higher than the leg temperature compensation can • d/p Cells
high level connection (tap) so that it be provided locally. Alternatively, Because the designs of the various
will maintain a constant condensate the host or supervisory control sys- d/p cells are discussed in detail in
level. Excess liquid will drain back tem can perform the compensation another issue of Transactions, only a
into the tank. It is also desirable calculations. brief overview is provided here.
either to install a level gage on the If it is desired to keep the process The motion balance cell is well
condensate pot or to use a sight flow vapors in the tank, a pressure repeater suited for remote locations where
indicator in place of the pot, so that can be used. These devices repeat the instrument air or electric power are
the level in the pot can conveniently vapor pressure (or vacuum) and send not available. If a bellows is used as
be inspected. out an air signal identical to that of the sensing element in a motion bal-
Either method (wet or dry) the vapor space. The measurement ance d/p cell, an increase in the pres-
assures a constant reference leg for side of the repeater is connected to sure on either side causes the corre-
the d/p cell, guaranteeing that the the vapor space and its output signal sponding bellows to contract (Figure
only variable will be the level in the to the low pressure side of the d/p 7-3A). The bellows is connected to a
TRANSACTIONS Volume 4 77
Pressure/Density Level Instrumentation 7
linkage assembly that converts the lin- a draft range of 0-H inH2O or as can accumulate and affect the
ear motion of the bellows into a wide as 0-1,000 psid. Some elec- performance of the cell. Flat and
rotary indicator motion, which can be tronic d/p cells can operate at line extended diaphragm-type d/p cells,
Bimetallic
Pneumatic
Range
Nozzle & Flapper Relay
Temperature
Liquid
Compensator Spring
Fill Feedback Bellows
Air
Supply
Output
Fulcrum & Seal
Force Bar
Low Pressure
High Pressure
Side Side
Figure 7-3:
Differential Pressure Cell Designs
calibrated to indicate the tank level. pressures up to 4,500 psig at 250°F. pressure repeaters, and chemical
In a force-balance type of d/p The drift and inaccuracy of some of seals are available to protect d/p
cell, the sensing element (often a these units have been tested for cells under these conditions.
diaphragm) does not move. A force periods of up to 30 months, and the Chemical seals, or diaphragm
bar is provided to maintain the errors did not exceed the ±0.5% of pressure seals, are available with fill
forces acting on the diaphragm in span limit. liquids such as water, glycol, alco-
equilibrium (Figure 7-3B). In pneumatic hol, and various oils. These seals are
d/p cells, this is often achieved by • Difficult Process Fluids used when plugging or corrosion
the use of a nozzle and flapper When the process fluid is a sludge, a can occur on both sides of the cell.
arrangement that guarantees that the viscous polymer or is otherwise hard A broad range of corrosion-resistant
pneumatic output signal will always to handle, the goal is to isolate the diaphragm and lining materials is
be proportional to the differential dirty process from the d/p cell. A available. Teflon® lining is often used
pressure across the cell. The output of flat diaphragm can be bolted to a to minimize material build-up and
pneumatic d/p cells is linear and is block valve on the tank nozzle so coating. Level measurement accuracy
usually ranged from 3 to 15 psig. The that the d/p cell can be removed for does suffer when these seals are
levels represented by such transmit- cleaning or replacement without tak- used. Capillary tube lengths should
ted signals (pneumatic, electronic, ing the tank out of service. If it is be as short as possible and the
fiberoptic or digital) can be displayed acceptable to take the tank out of tubes should be shielded from the
on local indicators or remote instru- service when d/p cell removal is sun. In addition, either low thermal
ments. Pneumatic transmitters require needed, an extended diaphragm expansion filling fluids should be
a compressed air (or nitrogen) supply. design can be considered. In this used or ambient temperature com-
Electronic d/p cells provide case, the diaphragm extension fills pensation should be provided, as
±0.5% of span or better precision the tank nozzle so that the discussed in connection with wet
typically conveyed via a 4-20 mA diaphragm is flush with the inside legs. If the seals leak, maintenance
signal. The range of these simple surface of the tank. This eliminates of these systems is usually done
and robust cells can be as narrow as dead ends or pockets where solids at the supplier’s factory due to
78 Volume 4 TRANSACTIONS
7 Pressure/Density Level Instrumentation
the complex evacuation and back- end of the dip pipe should be located a pressure at least 10 psi greater
filling procedures involved. far enough above the tank bottom so than the expected maximum total
that sediment or sludge will not plug pressure required (when the tank is
Bubbler Tubes it. Also, its tip should be notched full and the vapor pressure is at its
Bubbler tubes provide a simple and with a slot or “V” to ensure the for- maximum). An alternative to a con-
inexpensive but less accurate (±1-2%) mation of a uniform and continuous tinuous bubbler is to use a hand
level measurement system for corro- flow of small bubbles. An alternative pump (similar to a bicycle tire
sive or slurry-type applications. to locating the dip pipe in the tank is pump) providing purge air only
Bubblers use compressed air or an to place it in an external chamber when the level is being read.
inert gas (usually nitrogen) intro- connected to the tank. Bubblers do consume inert gases,
duced through a dip pipe (Figure 7-4A). In pressurized tanks, two sets of which can later accumulate and
Gas flow is regulated at a constant dip pipes are needed to measure blanket processing equipment. They
rate (usually at about 500 cc/min). A the level (Figure 7-4B). The two also require maintenance to ensure
differential pressure regulator across back-pressures on the two dip that the purge supply is always avail-
a rotameter maintains constant flow, pipes can be connected to the two able and that the system is properly
while the tank level determines the sides of a u-tube manometer, a dif- adjusted and calibrated. When all
back-pressure. As the level drops, the ferential pressure gage or a d/p factors are considered, d/p cells
L1
SS dPCV
F1 Remotely
Located
PCV Components
P1
N2
dPCV
SS F1
Spring
Tube
Diaphragm #1 Float
Equalizing Line
P2 Transmission Line
Regulator
Valve L1 Dip Tube
Flow Control
Manometer
Spring
Valve (V)
Flow @ P1
#2
Inlet Pressure
PNEUMATIC
SUPPLY
(N2)
A) Open Tank B) Closed Tank
TRANSACTIONS Volume 4 79
Pressure/Density Level Instrumentation 7
account for the difference in eleva- the high pressure side of the d/p cell saturated liquid layer (0.76) vary not
tion. This calibration adjustment is should be connected to the tank if only with drum pressure but also
called zero elevation when the cell is the specific gravity of the wet leg fill- with steaming rate. This causes the
located above the lower tap, and is ing fluid is close to that of the light swelling of bubbles when the
called zero suppression or zero layer. It should be connected to the steaming rate rises (and SG2 drops),
as well as their collapse when the
steaming rate drops (and SG2 rises).
Steaming
Therefore, to make an accurate
Rate
determination of both the level and
the mass of the water in the steam
drum, the calculation must consider
not only the d/p cell output, but
also the drum pressure and the pre-
vailing steaming rate.
h1
120°F
• Tank Farms
h3
Computerized tank farm systems usu-
525°F, 850 psig ally accept level signals from several
h2
80 Volume 4 TRANSACTIONS
7 Pressure/Density Level Instrumentation
the buoyant force acting on an object to 80°C (-40 to 180° F) and up to 150 (SG) of the process fluid. For clean liq-
equals the weight of the fluid dis- psig for rubber or plastic floats, and - uids a 0.1 SG difference might suffice,
placed. As the level changes around 40 to 260°C (-40 to 500°F) and up to while for viscous or dirty applications,
Switching Element
Pivot
Bias Spring Float Reed Switch
Magnet
B)
Float Permanent
1/2 NPT
Magnet
Level Differential (13 mm)
Rising Level
Falling Level
A) Magnetic Piston B) Reed Switch C) Mecury Switch
the stationary (and constant diameter) 750 psig for stainless steel floats. a difference of at least 0.3 SG is rec-
displacer float, the buoyant force Standard float sizes are available from ommended. This provides additional
varies in proportion and can be detect- 1 to 5 inches in diameter. Custom float force to overcome the resistance due
ed as an indication of level. Regular sizes, shapes, and materials can be to friction and material build-up. In
and displacer floats are available as ordered from most manufacturers. The dirty applications, floats should also
both continuous level transmitters and float of a side-mounted switch is hor- be accessible for cleaning.
point-sensing level switches. izontal; a permanent magnet actuates Floats can be attached to mechan-
In industrial applications, displacer the reed switch in it (Figure 7-6B). ical arms or levers and can actuate
floats are often favored because they Floats should always be lighter than electrical, pneumatic, or mechanical
do not require motion. Furthermore, the minimum expected specific gravity mechanisms. The switch itself can be
force can often be detected more
accurately than position. However,
regular floats are also used, mostly
for utilities and in other secondary
applications.
TRANSACTIONS Volume 4 81
Pressure/Density Level Instrumentation 7
mercury (Figures 7-6A and 7-6C), dry duplex sump-pump stations. A sim- and/or visual alarm.
contact (snap-action or reed type, plex (one pump) system will use a Figure 7-8A illustrates how a side-
shown in Figure 7-6B), hermetically single switch wired in series with the mounted float switch might actuate
sealed, or pneumatic. The switch can motor leads so that the switch an adjacent, sealed reed switch. The
main advantage of this design is that
Top Mounted the lever extension tends to amplify
the buoyant force generated by the
B)
C) float. Therefore the float itself can
be rather small. The main disadvan-
A) tage is that the tank must be opened
in order to perform maintenance on
the switch. If the buoyant force of
Side
Float
the float is used mechanically to
Mounted
Cage actuate a snap-action switch, a force
of only one ounce is needed.
In top (or bottom) mounted mag-
netic float switches (Figure 7-8B), the
magnet is in the cylindrical float that
travels up or down on a short vertical
guide tube containing a reed switch.
The float’s motion is restrained by
clips and can be only H in or less.
Figure 7-8: Typical Level Switch Applications These float and guide tubes are avail-
be used to actuate a visual display, directly starts and stops the pump able with multiple floats that can
annunciator, pump, or valve. The elec- motor (Figure 7-7). detect several levels. The switch
tric contacts can be rated light-duty (10- A duplex (two pump) application assembly itself can be either inserted
100 volt amps, VA) or heavy-duty (up to might use three switches: one at the directly into the tank or side-mount-
15 A @ 120 Vac). If the switch is to oper- tank bottom (LO) to stop both ed in a separate chamber.
ate a circuit with a greater load than the pumps, another in the middle (HI) to A magnetic piston operated
rating of the switch contacts, an inter- start one pump, and the last at the switch also can be mounted in an
posing relay needs to be inserted. If the top (HI-HI) to actuate the second external chamber (Figure 7-8C). As
switch is to be inserted in a 4-20 mA dc pump, as well as perhaps an audible the magnet slides up and down
circuit, gold-plated dry contacts should
be specified to ensure the required very
low contact resistance.
82 Volume 4 TRANSACTIONS
7 Pressure/Density Level Instrumentation
inside a non-magnetic tube, it oper- • Displacer Switches while a displacer switch can be tested
ates the mercury switch outside the Whereas a float usually follows the simply by lifting a suspension (Figure
tube. These switches are complete- liquid level, a displacer remains par- 7-10A). Displacer switches are avail-
ly sealed and well suited for heavy tially or completely submerged. As able with heavy-duty cages and
duty industrial applications up to shown in Figure 7-10A, the apparent flanges for applications up to 5000
900 psig and 400°C (750°F), meeting weight of the displacer is reduced as psig at 150°C (300°F), suitable for use
ASME code requirements. These it becomes covered by more liquid. on hydraulic accumulators, natural
Torque Tube
Torque
Flange Torque
Arm Rod
Torque
Arm
Torque
Block Tube
Testing
Cable
Limit
Stop
Knife
Edge Nozzle
High Level Flapper
Flexible
Displacer
Cable Displacers
Low Level
switches can be side, top, or cage When the weight drops below the gas receivers, high pressure scrub-
mounted (Figure 7-9) and can serve spring tension, the switch is actuat- bers, and hydrocarbon flash tanks.
both alarm and control functions on ed. Displacer switches are more reli-
steam drums, feedwater heaters, able than regular floats on turbu- • Continuous Level Displacers
condensate pots, gas/oil separators, lent, surging, frothy, or foamy appli- Displacers are popular as level trans-
receivers, and accumulators. Light- cations. Changing their settings is mitters and as local level controllers,
duty caged float switches are also easy because displacers can be particularly in the oil and petro-
available for service ratings up to moved anywhere along the suspen- chemical industries. However, they
250 psig at 200°C (400°F) and 400 sion cable (up to 50 ft). These switch- are not suited for slurry or sludge
psig at 40°C (100°F)—suitable for es are interchangeable between service because coating of the dis-
many boilers, condensate receivers, tanks because differences in process placer changes its volume and there-
flash tanks, day tanks, holding tanks, density can be accommodated fore its buoyant force. They are most
and dump valve controls. The cages by changing the tension of the sup- accurate and reliable for services
can be provided with level gages. port spring. involving clean liquids of constant
Multiple switches are available for Testing the proper functioning of a density. They should be temperature-
multiple-switching applications such regular float switch may require fill- compensated, particularly if varia-
as boiler level alarms and controls. ing the tank to the actuation level, tions in process temperature cause
TRANSACTIONS Volume 4 83
Pressure/Density Level Instrumentation 7
significant changes in the density of force is balanced by a spring, there is standpipes with level gages and iso-
the process fluid. some movement, while with a force- lating valves (Figure 7-11). This way it
When used as a level transmitter, balance detector, the displacer stays is possible to recalibrate or maintain
the displacer, which is always heavier in one position and only the level the displacer without interrupting
than the process fluid, is suspended over the displacer varies. the process.
from the torque arm. Its apparent Displacer units are available with
• Interface Applications
When measuring the interface
between a heavy liquid and a light
2" or Larger Nozzle Pressure Vessel liquid (such as oil on water), the top
connection of the displacer is
3/4" Gate Valve (Vent)
placed into the light and the bot-
or Plug 2 " Gate Valve
Gate Valve
tom connection into the heavy liq-
2" or 3" Standpipe uid layer. If the output of such a
Automatic Gage Cocks
or Tees transmitter is set to zero when the
3/4" Gate Valve (Vent)
or Plug chamber is full of the light liquid,
1 1/2" or Larger
1 1/2" or Larger Screwed
and to 100% when it is full with the
Gate Valve or Flange Connections heavy phase, the output will corre-
spond to the interface level.
3/4" Coupling-6000 LB
Naturally, when interface is being
Tapped One End Only
measured, it is essential that the
two connections of the displacer
Overlapping
chamber be located in the two differ-
Gage Glasses ent liquid layers and that the chamber
1 1/2" or Larger
always be flooded. Displacer diame-
Gate Valve
ter can be changed to match the dif-
Reducer to 3/4" Pipe
ference in liquid densities, and dis-
These Assemblies
3/4" Gate Valve (Drain) placer length can be set to match the
May Be Elbows vertical range of the level interface
variation.
Regular floats can also be used
for interface detection if the differ-
Figure 7-11: Installation to API Standards
ence in SG between the two process
weight causes an angular displace- both pneumatic and electronic out- liquids is more than 0.05. In such
ment of the torque tube (a torsion puts and can also be configured as applications, a float density is need-
spring, a frictionless pressure seal). local, self-contained controllers. ed that is greater than the lighter
This angular displacement is linearly When used in water service, a 100 liquid and less than the heavier liq-
proportional to the displacer’s cubic inch displacer will generate a uid. When so selected, the float will
weight (Figure 7-10B). buoyant force of 3.6 pounds. follow the interface level and, in
Standard displacer volume is 100 Therefore, standard torque tubes are clean services, provide acceptable
cubic inches and the most commonly calibrated for a force range of 0-3.6 performance.
used lengths are 14, 32, 48, and 60 in. lbf and thin-walled torque tubes for a
(Lengths up to 60 ft are available in 0-1.8 lbf range. • Continuous Level Floats
special designs.) In addition to For oil refineries and other Of the various float sensor designs
torque tubes, the buoyant force can processes that are operated continu- used for continuous level measure-
also be detected by other force sen- ously, the American Petroleum ment, the oldest and arguably most
sors, including springs and force-bal- Institute recommends (in API RP 550) accurate is the tape level gage (Figure
ance instruments. When the buoyant that displacers be installed in external 7-12A). In this design, a tape or cable
84 Volume 4 TRANSACTIONS
7 Pressure/Density Level Instrumentation
connects the float inside the tank to with a visual indicator, consisting of This tank level sensing method is
a gage board or an indicating take-up G-in triangular wafer elements. highly accurate, to ±0.02 in, and there-
reel mounted on the outside of the These elements flip over (from fore is ideal for precision inventory
tank. The float is guided up and down green to red, or any other color) management operations. The sensor is
the tank by guide wires or travels when the magnet in the float reaches available in lengths of 2-25 ft and can
inside a stilling well. These level indi- their level (Figure 7-12B). Alarm be inserted into the tank from the top
cators are used in remote, unattend- switches and transmitter options of the vessel through flanged,
ed, stand-alone applications, or they are available with similar magnetic screwed, or welded connections. For
can be provided with data transmis- coupling schemes (Figure 7-12C). In a the simultaneous measurement of
sion electronics for integration into similar design, a series of reed both interface and total level, a two-
plant-wide control systems. switches is located inside a stand- float system is available (Figure 7-12D).
To install the tape gage, an open- pipe. The change in output voltage A resistance temperature detector
ing is needed at the top of the tank as the individual reed switches are (RTD) is also available for temperature
and an anchor is required at its bot- closed by the rising magnet is mea- compensation. Like all other float
tom. When properly maintained, sured, giving an indication of level. level instruments, this design too is
tape gages are accurate to ±G in. It is The operation of magnetostrictive for clean liquids. Rating is up to 150°C
important to maintain the guide sensors is based on the Villari effect. (300° F) and 300 psig. The transmitter
wires under tension, clean and free In the magnetic waveguide-type con- output can be 4-20 mA dc analog or
of corrosion, and to make sure that tinuous level detector, the float (or fieldbus-compatible digital.
the tape never touches the protec- floats, when detecting interface) trav-
tive piping in which it travels. If this els concentrically up and down out- • Float Control Valves
is not done, the float can get stuck side a vertical pipe. Inside the pipe is Float-operated control valves com-
on the guide wires or the tape can a concentric waveguide made of a bine level measurement and level
get stuck to the pipe. (This can hap- magnetostrictive material. A low cur- control functions into a single level
pen if the level does not change for rent interrogation pulse is sent down regulator. While simple and inexpen-
long periods or if the tank farm is the waveguide, creating an electro- sive, they are limited to applications
TRANSACTIONS Volume 4 85
Pressure/Density Level Instrumentation 7
Yet, for simple and unattended basin) is to first let the level drop so (Kc = Qmax/h), expressed in units of
applications (like controlling the that the sinking of the float will fur- GPM/inch. The offset of a float regu-
make-up water supply into a cooling ther open the valve. The relationship lator is the distance (in inches)
tower basin or draining condensate between the maximum flow through between the center of the float range
from a trap), they are acceptable. It is a linear valve (Qmax) and the range in and the amount of elevation of the
important to understand that float liquid level (h) is called the propor- float required to deliver the flowrate
regulators are simple proportional tional sensitivity of the regulator demanded by the process. T
controllers: they are incapable of
holding level at a single setpoint.
What they can do is open or close a References & Further Reading
valve as the float travels through its • OMEGA Complete Flow and Level Measurement Handbook and
control range. Therefore, instead of a Encyclopedia®, OMEGA Press, 1995.
setpoint, regulators have a throttling • OMEGA Volume 29 Handbook & Encyclopedia, Purchasing Agents
range. If the range is narrow (floats Edition, OMEGA Press, 1995.
usually fully stroke their valve over a • Instrument Engineer’s Handbook, Bela G. Liptak, editor, CRC Press, 1995.
few inches of float travel), it gives the • Instrumentation for Process Measurement and Control, Third Edition, N.
impression of a constant level. A. Anderson, Chilton, 1980.
In fact, level will vary over the • Measurement and Control of Liquid Level, C. H. Cho, Instrument Society
throttling range because the only of America, 1982.
way for the regulator to increase the • Principles of Industrial Measurement for Control Applications, E. Smith,
feed flow (say into a cooling tower Instrument Society of America, 1984.
86 Volume 4 TRANSACTIONS
8
C
apacitance level detectors are charge. The storage capability of a material (that is, the less conductive it
also referred to as radio fre- capacitor is measured in farads. As is). Capacitance (C) is calculated as:
quency (RF) or admittance shown in Figure 8-1, the capacitor
level sensors. They operate in plates have an area (A) and are sepa- C = KA/D
the low MHz radio frequency range, rated by a gap (D) filled with a non-
measuring admittance of an alter- conducting material (dielectric) of If the area (A) of and the distance (D)
nating current (ac) circuit that varies dielectric constant (K). The dielectric between the plates of a capacitor
with level. Admittance is a measure constant of a vacuum is 1.0; the remain constant, capacitance will
vary only as a function of the dielec-
Table 6: Applications of Capicitance Level Sensors
tric constant of the substance filling
INDUSTRY MATERIALS SENSED
Chemical/Petrochemical
the gap between the plates. If a
Soda Ash, Fuel, Oil , Clay, Liquids & Powders
change in level causes a change in the
Feed & Grain
Pellets, Granules, Flakes, Fats, Molasses, Calcium Dust
Food
total dielectric of the capacitance
Sugar, Salt, Flour, Powdered Milk, Various Liquids
system, because (as illustrated in
Pet Food
Pellets, Rawhides, Grains
Figure 8-1B) the lower part of area (A)
Plastics/Rubber
Plastic Pellets, Resin, Regrind, Powders, Rubber
is exposed to a liquid (dielectric Kl)
Foundries
Silica Sand, Foundry Sand
while the upper part is in contact with
Beer/Breweries
Malt, Barley Liquids
a vapor (dielectric Kv, which is close to
Pharmaceuticals
Various Powders & Liquids
1.0), the capacitance measurement
Power/Utilities
Coal, Wood, Sawdust, Petro-Coke
will be proportional to level.
Water/Waste Treatment
Limestone, Hydrated Lime, Water
In the case of a horizontally
Charcoal
Charred Sawdust, Wood
mounted level switch (Figure 8-2), a
Saw Mills/Woodworking
Wood Shavings, Sawdust
conductive probe forms one of the
Mining & Miscellaneous
Various Minerals, Clay, Metals, Stone, Glass, Bentonite
plates of the capacitor (A1), and the
of the conductivity in an ac circuit, dielectric constants of a variety of vessel wall (assuming it is made from
and is the reciprocal of impedance. materials are listed in Table 7. a conductive material) forms the
Admittance and impedance in an ac The dielectric constant of a sub- other (A2). An insulator with a low
circuit are similar to conductance stance is proportional to its admit- dielectric constant is used to isolate
and resistance in a direct current (dc) tance. The lower the dielectric con- the conductive probe from the hous-
circuit. In this chapter, the term stant, the lower the admittance of the ing, which is connected to the vessel
capacitance level sensor will be used
instead of RF or admittance. Level Kv
Table 6 lists some of the industries D #1 #2 C= KA
D
and applications where capacitance- - --
- Kl C=Capacitance
type level sensors are used. - -- K=Dieletric Constant
- -- --
- + ++ A=Area of Plates
- A
- -- ++ D=Dist. Between Plates
Theory of Operation --
-- + ++
++
++
++
Voltmeter Electron
A capacitor consists of two conduc- -- ++ A Flow
- ++ ++
++ + RF
tors (plates) that are electrically iso- ++ Ammeter
lated from one another by a non- ++
+
conductor (dielectric). When the
two conductors are at different A) Capacitor B) Capacitance Circuit
potentials (voltages), the system is
Figure 8-1: Capacitance Theory of Operation
capable of storing an electric
TRANSACTIONS Volume 4 87
RF/Capacitance Level Instrumentation 8
wall. The probe is connected to the longer surrounded by vapors (K1), but detectable change in capacitance
level sensor via the conductive threads by the process material (K2), the result- resulting from a change in dielectric
of the housing. Measurement is made ing capacitance change is directly constant (K2 -K1).
In most level-sensing applications,
the reference material is air (K1 = 1.0).
D K1 Table 7 gives the K2 values of a variety
of process materials. As the dielectric
-- ++
constant of the process material gets
-- ++ A1 close to that of air (K2 for plastic pel-
lets, for example, is 1.1), the measure-
A2 Insulation
ment becomes more difficult.
K2
Probe Designs
-- ++ The most common probe design is a
-- ++ stainless steel rod of G in. or H in.
diameter, suitable for most non-con-
-- ++ ductive and non-corrosive materials.
-- ++
The probe is insulated from the
A = A1 + A2 ∆C = (K2 + K1) A/D housing and bin wall by an low-
dielectric insulator, such as Nylon or
Figure 8-2: Horizontally Mounted Capacitance Switch
Ryton. These polymers have maxi-
by applying an RF signal between the related to the difference in dielectric mum operating temperatures of 175-
conductive probe and the vessel wall. constant between the two media: 230°C (350-450°F). Ceramics can be
The RF signal results in a minute used for higher temperature applica-
current flow through the dielectric Change in C = (K2 - K1)(A/D) tions or if abrasion resistance is
process material in the tank from the required. For applications where the
probe to the vessel wall. When the The sensitivity of a capacitance process material is conductive and
level in the tank drops and the probe sensor is expressed in pico-farads (pF). corrosive, the probe must be coated
is exposed to the even less conduc- The capacitance unit is the farad, with Teflon® or Kynar.
tive vapors, the dielectric constant defined as the potential created when Some point level sensors are avail-
drops. This causes a drop in the a one-volt battery connected to a able with build-up immunity, or coat-
capacitance reading and a minute capacitor causes the storage of one ing rejection functionality. This is
drop in current flow. This change is
detected by the level switch’s inter- Coating Vessel Wall
Driven
nal circuitry and translated into a Shield (+5 V)
change in the relay state of the level
switch. In the case of continuous
level detectors (vertical probes), the
output is not a relay state, but a Current Path
Insulation Measuring
scaled analog signal. Through Level
Section (+5 V)
The total area is the combined No Current Flow
area of the level sensor probe and
the area of the conductive vessel
wall (A = A1 + A2), and the distance (D) Figure 8-3: Coating Rejection Design
is the shortest distance between the coulomb of electric energy. A pico- required when the process material is
sensor probe and the vessel wall. farad is one trillionth of that, and the wet or sticky and likely to cause
Both of these values are fixed. sensitivity of an accurate capacitance permanent coating. Build-up immu-
Therefore, when the probe is no detector is 0.5 pF. This is the minimum nity is provided by the addition of a
88 Volume 4 TRANSACTIONS
8 RF/Capacitance Level Instrumentation
second active section of probe and a used to suspend the probe up to 15 m • Electronics & Housings
second insulator (Figure 8-3). This sec- (50 ft) (Figure 8-4B). Most capacitance The electronic circuitry of the probe
ond active section (the driven shield) level sensors are provided with I to performs the functions of: 1) rectifying
is driven at the same potential and 1-H in NPT mounting connectors. The and filtering the incoming power, 2)
frequency as the measuring probe. matching female coupling is usually generating the radio frequency signal,
Because current cannot flow between welded to the vessel wall and the 3) measuring the changes in current
equal potentials, the measuring probe capacitance probe is screwed into the flow, and 4) driving and controlling
interface devices such as relays,
analog signal generators and display
meters. The circuitry is usually of solid
state design and provided with poten-
tiometer adjustments for setting sen-
sitivity and time delays.
Because the level sensor will ulti-
mately drive an external device, it is
advisable to evaluate for system
Capacitance probes typically are coated with Teflon® (shown), Kynar, or polyethylene compatibility the number of relays
required, their capacities, or the
does not sense material build-up mating connector. Low profile capac- analog signals required, time delays,
between the probe and vessel wall. itance sensors also are available and power supply requirements.
Typical insertion lengths of stan- (Figure 8-4C) and are flange-mounted. More advanced microprocessor-
dard capacitance probes range from In applications where the vessel is based units are self-calibrating; sen-
7 to 16 in. These probes typically are non-conductive and unable to form sitivity and time delay adjustments
side-mounted (Figure 8-4A). Vertical the return path for the RF signal, a are under pushbutton control.
probes can be extended by solid rods second probe placed parallel to the These units are often supplied with
up to a length of 1.2 to 1.5 m (4 to 5 ft), active one or a conductive strip can self-test capability and built-in tem-
or a steel cable with a weight can be be installed. perature compensation.
1-14" NPT
25 mm
(1")
D
Pipe Section Steel Plates 115 mm (4.5")
175 mm
(7") Flexible Cable 15 m (50') Max.
Customer
D 95 mm
Specified
(3.75") Length 50 mm
(2") Flange
Cable Weight
200 mm (7.8")
3
4"
NPT
TRANSACTIONS Volume 4 89
RF/Capacitance Level Instrumentation 8
The more advanced designs are with time delays for filtering out • The Dielectric Constant
also two-wire, intrinsically safe, and false readings caused by material The dielectric constant of the process
supply your choice of standard 4-20 shifts or splashing liquids. In addition, material is the most important aspect
mA or digitally enhanced output the feature of failsafe selectability of the process data. The higher the
using the HART (Highway provides a predetermined state for difference between the dielectric
constants (of the process material and
Table 7: Dielectric Constants
the vapor space or between the two
SOLIDS
DIELECTRIC CONSTANT
DIELECTRIC CONSTANT
layers in the case of an interface mea-
Acetic Acid
4.1
Phenol
4.3
surement), the easier the measure-
Asbestos
4.8
Polyethylene
4.5
ment. If the difference is low (K2-K1 <
Asphalt
2.7
Polypropylene
1.5
1.0 in Figure 8-2), a high sensitivity
Bakelite
5.0
Porcelain
5.7
Calcium Carbonate
9.1
Quartz
4.3
design (0.5 pF) must be used.
Cellulose
3.9
Rubber (Hard)
3.0
Each sensor has a capacitance
Ferrous Oxide
14.2
Sand
3.5
Glass
3.7
Sulphur
3.4
threshold, defined as the amount of
Lead Oxide
25.9
Sugar
3.0
capacitance change required to
Magnesium Oxide
9.7
Urea
3.5
cause a change in the sensor output.
Naphthalene
2.5
Zinc Sulfide
8.2
Nylon
45.0
Teflon®
2.0
The dielectric constant of a material
Paper
2.0
can change due to variations in tem-
LIQUIDS
perature, moisture, humidity, materi-
TEMP
DIELECTRIC
TEMP
DIELECTRIC
°F/°C
CONSTANT
°F/°C
CONSTANT
al bulk density, and particle size. If
Acetone
71/22
21.4
Heptane
68/20
1.9
the change in dielectric constant
Ammonia
-27/-33
22.4
Hexane
68/20
1.9
results in a greater capacitance
Aniline
32/0
7.8
Hydrogen Chloride
87/28
4.6
change than the calibrated capaci-
Benzene
68/20
2.3
Iodine
224/107
118.0
Benzil
202/94
13.0
Kerosene
70/21
1.8
tance threshold of the sensor, a false
Bromine
68/20
3.1
Methanol
77/25
33.6
reading will result. This condition can
Butane
30/-1
1.4
Methyl Alcohol
68/20
33.1
Carbon Tetrachloride
68/20
2.2
Methyl Ether
78/26
5.0
usually be corrected by reducing the
Castor Oil
60/16
4.7
Mineral Oil
80/27
2.1
sensitivity (increasing the capaci-
Chlorine
32/0
2.0
Naphthalene
68/20
2.5
tance threshold) of the sensor.
Chloroform
32/0
5.5
Octane
68/20
2.0
Cumene
68/20
2.4
Pentane
68/20
1.8
As shown in connection with
Cyclohexane
68/20
2.0
Phenol
118/47
9.9
Figure 8-3, sensitivity can be
Dimethylheptane
68/20
1.9
Phosgene
32/0
4.7
increased by increasing the probe
Dimethylpentane
68/20
1.9
Propane
32/0
1.6
Dowtherm
70/21
3.3
Pyridine
68/20
12.5
length (A) or by decreasing the size
Ethanol
77/25
24.3
Styrene
77/25
2.4
of the gap (D). Either or both changes
Ethyl Acetate
68/20
6.4
Sulphur
752/400
3.4
will minimize the effect of dielectric
Ethyl Benzene
68/20
2.5
Toluene
68/20
2.4
Ethyl Benzene
76/24
3.0
Urethane
74/23
3.2
constant fluctuations or increase
Ethyl Ether
68/20
4.3
Vinyl Ether
68/20
3.9
sensitivity to low dielectrics. It is
Ethylene Chloride
68/20
10.5
Water
32/0
88.0
usually more practical to specify a
Formic Acid
60/16
58.5
Water
68/20
80.0
Freon 12
70/21
2.4
Water
212/100
48.0
longer probe than to decrease the
Glycol
68/20
41.2
Xylene 68/20 2.4 distance (D) from the vessel wall.
When the probe is installed from the
Addressable Remote Transducer) the relay output in the event of a side (Figure 8-4A), D is fixed, whereas if
protocol. Accuracy (including linearity, power failure or malfunction. the probe is inserted from the top of
hysteresis, and repeatability, but Sensor housings are typically the tank, D can be changed (if other
excluding temperature and supply made from cast aluminum, steel, or considerations permit) by moving the
voltage effects) is typically 0.25% of synthetic materials such as glass- probe closer to the wall of the vessel.
range. Minimum span is 4 pF, and the reinforced nylon. Most housings are If the same vessel will hold differ-
upper range limit (URL) is 2,500 pF. suitable for outdoor installations in ent materials at different times, the
Level switches are usually provided dusty or wet environments. capacitance sensor must be
90 Volume 4 TRANSACTIONS
8 RF/Capacitance Level Instrumentation
B • Application Considerations
Materials that are conductive (water-
based liquids with a conductivity of
100 micromhos/cm or more) can
cause a short circuit between a bare
Probes
Probes
Do Not
Touch the
stainless steel probe and the vessel
Touch the
Vessel Structure wall. As the liquid level drops, the
Vessel Structure
probe remains wetted, providing a
C conductive path between the probe
and the vessel wall. The faster the level
Complete Probe
changes, the more likely this false indi-
Protruding into Material
cation is to occur. It is advisable to use
Teflon® or Kynar insulator coating on
the conductive probe surface when
Portion of Probe
the process fluid is conductive.
Within Mounting Neck Temperature affects both the sen-
sor components inside the vessel
(active probes and insulators) and the
D
electronic components and housing
outside. An active probe is typically
made from stainless steel and, as
such (unless it is coated), it is suitable
Material
Material Can
for most applications. Probe insula-
Will Fall Away Build Up, Creating
tors can be Teflon®, Kynar, or ceramic,
A False Signal
and should be selected for the oper-
E ating temperature of the application.
The housing and the electronics are
affected by both the internal and
external vessel temperatures.
Ambient temperature limits usually
are specified by the manufacturer,
but heat conduction from a high-
Sensors Acceptable (> 18") Sensors Too Close (< 18") temperature process is more difficult
to evaluate. Heat conduction can be
reduced by using an extended
Figure 8-5: Capacitance Probe Installation Recommendations
mounting coupling or one made of a
TRANSACTIONS Volume 4 91
RF/Capacitance Level Instrumentation 8
low thermal conductivity material. If the probe and insulator. When mea- bridge between the active probe and
such methods are insufficient, the suring low-dielectric materials, it's the vessel wall. In addition, the probe
electronics may be mounted up to important that the entire probe be should not be mounted at an upward
20 ft away and connected via coaxial covered, not just the tip (Figure 8- angle (Figure 8-5D), to avoid material
cable. The cable’s inherent capaci- 5C). When rod or cable extensions build-up.
tance, however, reduces the overall are used, allow for 8-12 in. of active If more than one capacitance level
sensitivity of the system. probe coverage. sensor is mounted in the vessel, a min-
Housings must also be compati- Install the probe so that it does imum distance of 18 in. should be pro-
ble with the requirements for haz- not contact the vessel wall (Figure 8- vided between the probes (Figure 8-5E).
ardous, wash-down, wet, and/or 5B) or any structural elements of the Closer than that and their electro-
dusty environments. Explosion- vessel. If a cable extension is used, magnetic fields might interfere. If a
proof environments may require the allow for swinging of the cable as the capacitance probe is installed through
housing to be certified. In addition, material level in the vessel rises, so the side wall of a vessel and the
the active probe might need to be that the plumb bob on the end of weight of the process material acting
intrinsically safe. the cable does not touch the vessel on the probe is sometimes excessive,
If the process material is corrosive wall. The probe should not be a protective baffle should be installed
to stainless steel, the probe should mounted where material can form a above the sensor (Figure 8-4A). T
be coated with Kynar or Teflon® for
protection. Ryton is a good choice
for abrasive materials, and, for food References & Further Reading
grade or sanitary applications, stain- • OMEGA Complete Flow and Level Measurement Handbook and
less steel and Teflon® are a good Encyclopedia®, OMEGA Press, 1995.
probe-insulator combination. • OMEGA Volume 29 Handbook & Encyclopedia, Purchasing Agents
Edition, OMEGA Press, 1995.
Installation Considerations • Instrument Engineer’s Handbook, Bela Liptak, Third Edition, CRC Press, 1995.
The capacitance probe should be • Instrumentation for Process Measurement and Control, Third Edition, N.
mounted in such a way that its A. Anderson, Chilton, 1980.
operation is unaffected by incom- • Measurement and Control of Liquid Level, C. H. Cho, Instrument Society
ing or outgoing material flow of America, 1982.
(Figure 8-5A). Material impacts can • Principles of Industrial Measurement for Control Applications, E. Smith,
cause false readings or damage to Instrument Society of America, 1984.
92 Volume 4 TRANSACTIONS
9
A
n entire class of level instru- microwaves range from 1-300 GHz) and level. Time of flight is the period
mentation devices is based by their power levels (radar is around between the transmission of the
on a material’s tendency to 0.01 mW/cm2, while microwaves range radar pulse and the reception of the
reflect or absorb radiation. from 0.1-5 mW/cm2). Because return echo. It is determined by the
For continuous level gages, the most microwaves operate at a higher energy radar detector, which is simultane-
common types of radiation used are level, they can withstand more coating ously exposed to both the sent and
radar/microwave, ultrasonic, and than can radar-type sensors. the reflected signal. The detector
nuclear. Optical electromagnetic Radar sensors consist of a trans- output is based on the difference. The
radiation also can be used, but this mitter, an antenna, a receiver with frequency-modulated (FM) signal
has found its way primarily into the
point-switch applications discussed
in the next chapter. Large Size
Small Size
Datum Point
The main advantage of a radia- Parabolic Dish
Horn
Antenna Antenna
tion-based level gage is the absence
of moving parts and the ability to
detect level without making physi-
E d
cal contact with the process fluid.
Because they can in effect “see”
through solid tank walls, nuclear
radiation gages are perhaps the ulti-
mate in non-contact sensing. L
Because they require a gamma radi-
ation source and are relatively
expensive, however, nuclear gages A) B)
are often considered the level gage
of last resort. Figure 9-1: Electronics Calculate Time of Flight
Radar & Microwave signal processor, and an operator varies from 0 to 200 Hz as the dis-
In 1925, A. Hoyt Taylor and Leo Young interface. The transmitter is mounted tance to the process fluid surface
of the U.S. Navy used radar (RAdio on top of the vessel. Its solid-state varies between 0 and 200 ft. Because
Detection And Ranging) to measure oscillator sends out an electromag- this measurement takes place in the
the height of the earth’s ionosphere. netic wave (using a selected carrier frequency domain, it is reasonably
By 1934, they were developing radar frequency and waveform) aimed free of noise interference.
for Navy ships. In 1935, Robert downward at the surface of the The depth of the vapor space (the
Watson-Watt of England used radar process fluid in the tank. The fre- distance between the datum point
to detect aircraft. The first radar level quency used is typically 10 GHz. and the level in the tank, identified as
sensors were introduced in 1976, but The signal is radiated by a parabolic “d” in Figure 9-1B) is calculated from
they did not become economically dish or horn-type antenna (Figure 9- the time of flight (t) and the speed of
competitive until a decade later. 1A) toward the surface of the process light (c = 186,000 miles/sec):
Both radar signals and microwaves liquid (Figure 1B). A portion is reflect-
travel at the speed of light, but are ed back to the antenna, where it is d = t/2c
distinguished by their frequencies collected and routed to the receiver.
(FM radio broadcast frequency Here, a microprocessor calculates The level (L in Figure 9-1B) is calculated
is from 88 to 108 MHz, while the time of flight and calculates the by figuring the difference between
TRANSACTIONS Volume 4 93
Radiation-Based Level Instrumentation 9
the total tank height (E) and the • Contact & Non-Contact Radar to be partially reflected. The time-of-
vapor space depth (d): Non-contact radar gages either use flight is then measured (Figure 9-2B).
pulsed radar waves or frequency- The unreflected portion travels on to
L = E-d modulated continuous waves the end of the probe and provides a
(FMCW). In the first, short-duration zero-level reference signal. Contact
Knowing the signal velocity (c) and the radar pulses are transmitted and the radar technology can be used on liq-
dielectric constant (dc) of the vapor target distance is calculated using uids and on small-grained bulk solids
(that is, the relative ability of the vapor the transit time. The FMCW sensor with up to 20-mm grain size.
to oppose and reflect electromagnetic sends out continuous frequency- Reflection-type microwave switch-
waves), the velocity of the radar wave modulated signals, usually in succes- es measure the change in amplitude
transmission (V) can be calculated: sive (linear) ramps. The frequency of a reflected signal (Figure 9-3A). Air
V = c/(dc)0.5
Cutaway View of
Cutaway View of
Parabolic Dish
Horn Antenna
Antenna
• Antenna Designs and Mounting
Impulse
The two commonly used antennas are
the horn and the parabolic dish anten-
na. When the radar level gage sends
out its signal, the microwaves spread
out. The larger the antenna diameter, Process
Process
the smaller the divergence angle and Seal
Seal
Tank Nozzle
the greater the signal strength (Figure
9-1A). The disadvantages of smaller A) Non-Contact Tank Ceiling B) Contact
antennas include higher beam spread-
ing and the correspondingly increased
Figure 9-2: Radar-Based Antenna Installations
possibility of reflection from obsta-
cles within the tank. On the positive difference caused by the time delay and vapors return a small percentage
side, there is a greater chance that the between transmittal and reception of the signal because of their low
emitted beam will be reflected back indicates the distance. dielectric constants, while high
to the detector. Therefore, alignment Radar beams can penetrate plas- dielectric materials such as water
of the sensor is not as critical. tic and fiberglass; therefore, non- return almost all the signal. More sen-
Large antennas generate a more contact radar gages can be isolated sitive switches can distinguish liquid-
focused signal, helping to eliminate from the process vapors by a seal. liquid or liquid-solid interfaces having
noise interference from flat and hor- The seal can be above the parabolic as little as 0.1 difference in dielectric
izontal metal surfaces. On the other disc (Figure 9-1A) or can totally iso- constant. Low dielectric materials like
hand, they are more prone to errors late the sensor (Figure 9-2A). The plastic pellets (dielectric 1.1) can be
caused by unwanted reflections from beam’s low power allows for safe measured if the particle diameter is
turbulent or sloping surfaces. A fully installation in both metallic and less than 0.1 in (larger than that, exces-
isolated antenna mounted outside non-metallic vessels. Radar sensors sive beam scattering occurs).
the tank (Figures 9-2 and 9-4) pro- can be used when the process mate- The beam-breaker switch sends a
vides both sealing and thermal isola- rials are flammable or dirty and microwave beam from a transmitter
tion. If the antenna is positioned when the composition or tempera- to a receiver located on the opposite
below the process seal, it is exposed ture of the vapor space varies. side of the tank. When the beam is
to the process vapors, but gains the Contact radar gages send a pulse blocked, the signal is weakened
advantages of stronger signal ampli- down a wire to the vapor-liquid inter- (Figure 9-3B). Beam-breaker align-
tudes and suitability for higher oper- face. There, a sudden change in the ment is not critical, and separation
ating pressures. dielectric constant causes the signal distance can be up to 100 ft.
94 Volume 4 TRANSACTIONS
9 Radiation-Based Level Instrumentation
Both reflection and beam-breaker advantages. For example, ultrasonic sound in atmospheric air is 340 m/s or
microwave switches are typically used sensors are affected by the composi- 762 mph. At that same temperature, an
in applications where it is desirable tion of the vapor space. On the other ultrasonic pulse travels through water
not to penetrate the tank. These non- hand, ultrasonic sensors perform at 1,496 m/s or 3,353 mph. If the air is
intrusive sensors send electromagnet- better in dirty applications, or with heated to 100°C, the speed of sound
ic radio waves through plastic, ceramic solids when the grain size is larger rises to 386 m/s. Indeed, the speed of
or glass windows, or through fiber- than 20 mm. sound is proportional to the square
glass or plastic tank walls. root of temperature. At near ambient
Ultrasonic Level Gages temperatures, the speed rises by 0.6
• Advantages & Limitations The origin of ultrasonic level instru- m/s per each 1°C increase, correspond-
The reflective properties of the mentation goes back to the echome- ing to an increase of 0.18%/°C.
process material affect the returned ters used in measuring the depth of Ultrasonic level switches (point
radar signal strength. Whereas liquids wells by firing a blank shell and tim- sensors) operate by detecting either
have good reflectivity characteristics, ing the return of the echo. SONAR dampening of ultrasonic oscillation or
solids do not. Radar can detect the detectors used in naval navigation by sensing the absorption or transmis-
liquid level under a layer of light dust also predate industrial applications sion of an ultrasonic pulse. Ultrasonic
or airy foam, but if the dust particle of this principle. level transmitters measure actual dis-
size increases, or if the foam or dust The frequency range of audible tance by issuing an ultrasonic pulse
gets thick, it will no longer detect the sound is 9-10 kHz, slightly below the and measuring the time required for
liquid level. Instead, the level of the 20-45 kHz range used by industrial the reflected echo to be received.
foam or dust will be measured. level gages. The velocity of an ultra-
Internal piping, deposits on the sonic pulse varies with both the sub- • Ultrasonic Transducers
antenna, multiple reflections, or stance through which it travels and The transducer that generates the
reflections from the wall can all inter- with the temperature of that sub- ultrasonic pulse is usually piezoelec-
fere with the proper operation of a stance. This means that if the speed tric, although in the past electrostatic
radar sensor. Other sources of inter- of sound is to be used in measuring a units also were used. An electrostatic
ference are rat-holing and bridging of level (distance or position), the sub- transducer is constructed of a thin,
solids, as well as angled process stance through which it travels must flexible gold-plated plastic foil,
Reflection
Microwave
Detector Microwave
Microwave
Reflected
Transmitted
Transmitter Beam Receiver
Beam
Absorbed
Microwave
Beam Microwave
Microwave
Window Window
Window
A) Reflection B) Beam-Breaker
Figure 9-3: Microwave Switches Send Pulses Through a Window or Tank Wall
material surfaces that can reflect the be well known and its temperature stretched over an aluminum back-
radar beam away from the receiver. variations must be measured and plate and held in place by a leaf
In comparison to other radiation compensated for. spring. This design was used in early
reflection sensors, radar has some At room temperature, the speed of Polaroid auto-focus cameras and is
TRANSACTIONS Volume 4 95
Radiation-Based Level Instrumentation 9
still utilized in clean environments. receiver elements (Figure 9-4A). Most determine the volume of liquid.
Piezoelectric transducers utilize often, however, a single transducer is If it is desired to measure the
Ground Level
Casing
Brine Pipe
Hydrocarbon
Cavity
Interface
Brine
Transducer
ceramic or polymer crystals vibrated cycled on and off at regular intervals height of the liquid column directly,
at their natural frequency. These to listen for the reflected echo (Figure the transducer can be mounted in
units are much more rugged, can 9-4A). When mounted on the top of the bottom of the tank (Figure 9-4A).
withstand wash-down pressures of the tank, the sensor detects the depth However, this configuration exposes
1,200 psig and can conform to of the vapor space. Accurate knowl- the transducer to the process fluid
NEMA-6P (IEC IP67) standards. and limits accessibility for mainte-
Generally, the larger the diameter nance. Alternately, the transducer
of the transducer, the longer the can be mounted on the outside of
range and the lower the frequency. the wall of the vessel bottom, but
This is because, after releasing an the ultrasonic pulse is likely to be
ultrasonic pulse, the transducer needs substantially weakened by the
time for the vibration to settle. The absorbing and dispersing effects of
oscillation frequency is inversely pro- the tank wall (Figure 9-4A).
portional to the element’s diameter, Stagnant, unagitated liquids and
so smaller diameter transducer ele- solids consisting of large and hard
ments generate higher frequencies. particles are good reflectors, and
Standard transducers have a beam therefore good candidates for ultra-
angle of about 8°, require a connec- sonic level measurement. Fluff,
tion size between G in and 2.5 in NPT, foam, and loose dirt are poor reflec-
and are suited for operating tempera- tors, and dust, mist, or humidity in
tures in the range of -20 to 60°C (-30 the vapor space tend to absorb the
to 140°F). Accuracy is typically within ultrasonic pulse. The ultrasonic sig-
0.25-0.5% of full range, up to about 30 nal also is attenuated by distance. If
ft. Output typically is 4-20 mA with a a 44-kHz sound wave is traveling in
12-amp relay output. dry, clean ambient air, its sound
power drops by 1-3 decibels (dB) for
• Level Transmitter Configurations Figure 9-5: Calibration Target each meter of distance traveled.
The ultrasonic level sensor assembly edge of the shape of the tank’s cross- Therefore it is important, particularly
can consist of separate transmitter and section is required in order to when measuring greater depths,
96 Volume 4 TRANSACTIONS
9 Radiation-Based Level Instrumentation
that the transducers generate a time it takes for the echo to return which, through multiplexing, can
strong and well-focused ultrasonic is an indication of the location of reduce the unit costs of obtaining
pulse (Figure 9-4B). the interface (Figure 9-4C). level measurements.
It is also desirable that the surface
be both flat and perpendicular to the • Special Features • Level Switches
sound wave. In liquid-level applica- Most modern ultrasonic instruments When it is sufficient to detect the pres-
tions, the aiming angle must be with- include temperature compensation, ence or absence of level at a particular
in 2 degrees of the vertical. If the sur- filters for data processing and elevation, dampened or absorption-
face is agitated or sloping (as in the response time, and some even pro- type level switches can be considered.
case of solids), the echo is likely to vide self-calibration. Figure 9-5 illus- In the dampened design, a piezoelec-
be dispersed. Therefore, the key to trates a fixed target assembly that tric crystal vibrates the sensor face at
100
80
60
50
SG=0.5
40
30
% Transmission
20
Water
SG=1.0
10
8
6 Steel
5 SG=7.6 SG=1.5
4
3 Aluminum
SG=2.0
SG=2.8
2
Lead
SG=11.3
1
1 2 3 4 5 6 7 8 9 10 12 14 16 18 20 22 24 26 28 30
Thickness - Inches
successful ultrasonic level sensor provides a point reference to auto- its resonant frequency. The vibration is
installations is the careful analysis of matically recalibrate the level sensor. dampened when the probe face is sub-
the reflection, propagation, and Multiple calibration targets can be merged in process fluid. As shown in
absorption characteristics of the provided by calibration ridges in Figure 9-3A, these switches can be
tank’s contents. sounding pipes. This can guarantee mounted outside or inside the tank,
When detecting the interface measurement accuracy of within above or below the liquid level. The
between two liquids, such as the 5 mm over a distance of 30 meters. probe can be horizontal or vertical.
hydrocarbon/brine interface in a Intelligent units can perform These switches are limited to clean liq-
salt dome storage well, the trans- automatic self-calibration or convert uid installations because coating can
ducer is lowered down to the bot- the level in spherical, irregular, or hor- dampen the vibration. Solids may not
tom of the well. The ultrasonic izontal cylindrical tanks into actual provide sufficient dampening effects to
pulse is sent up through the heavy volume. They can also be used in actuate the switch.
brine layer to the interface. The multi-tank or multi-silo installations, In the absorption-type level switch,
TRANSACTIONS Volume 4 97
Radiation-Based Level Instrumentation 9
one piezoelectric crystal serves as a current outputs are also used. a seemingly solid, impenetrable mass
transmitter and another as the receiver. The presence or absence of an of matter. In the passage, however,
When the gap between them is filled interface between clean liquids can the gamma rays lost some of their
with liquid, the sonic wave passes be measured by inserting an absorp- intensity. The rays were predictably
from one crystal to the other. When tion (gap) probe at a 10° angle below affected by the specific gravity and
vapors fill the gap, however, the ultra- the horizontal. In this configuration, as total thickness of the object, and by
sonic pulse does not reach the long as the probe is immersed in the the distance between the gamma ray
source and the detector.
For example, Figure 9-6 shows
Nonionizing Ionizing that, if radiation from Cesium 137 is
Photon Energy eV passing through an 3-in thick steel
object, 92% of the radiation energy
10-10 10-5 100 105 1010
will be absorbed and only 8% will be
transmitted. Therefore, if the observ-
Ultraviolet er can hold all variables except thick-
Electric
Waves Cosmic
ness constant, the amount of gamma
Radio Waves Infrared Gamma
Rays transmission can be used to measure
Rays the thickness of the object. Assuming
Visible X-Rays that the distance between the source
and detector does not change, one
can make accurate measurements of
either thickness (level), or, if thick-
105 100
10-5 10-10 10-15 ness is fixed, then of the density of a
Wavelength cm process material.
98 Volume 4 TRANSACTIONS
9 Radiation-Based Level Instrumentation
decays, it loses strength—the time it above the liquid, as the level rises in the This can be illustrated by an example:
takes to lose half of its strength is tank, the intensity at the detector
called its half-life. drops. When the tank is full, radiation • Source Sizing
The half-life of Cobalt 60 is 5.3 intensity is practically zero. A point source of 10 mCi Cesium 137
years. This means that, in 5.3 years, the When used as a tank level sensor, (source constant for Cesium 137 is
activity of a 100 millicurie (mCi) radiation must pass through several lay- K=0.6) is installed on a high-pressure
Cobalt 60 source will be reduced to ers of material before reaching the water tank having H-in steel walls
50 mCi. (One mCi is defined as the detector. At the detector, the maximum (Figure 9-9). Usually, two criteria need
rate of activity of one milligram of radiation must be less than some safety to be satisfied: First, the radiation
Radium 226.) When used for level limit (such as 5 mr/hr) to avoid the need intensity at the detector must drop by
measurement, the continuous loss of for “posting.” Other criteria can be used, at least 50% as the level rises from 0-
source strength requires not only con- such as keeping a yearly dosage under 5 100%. The second and more important
tinuous compensation, but, eventually rems (roentgen + equivalent + man). If criterion is that the maximum radiation
(in the case of Cobalt 60, in about 5 somebody is exposed to radiation dose at the detector (when the tank is
years), the source must be replaced. throughout the year, such a dosage will empty) must not exceed the safety
This means not only the expense of result from exposure to radiation at an limit (say, 2.4 mr/hr). It must exceed 1.0
purchasing a new source, but also the intensity of 0.57 mr/hr, while if an oper- mr/hr, however, in order to actuate the
cost of disposing of the old one. ator is exposed for only 40 hrs/wk, 5 intended ion chamber detector.
In contrast, the 33-year half-life of rem/yr will correspond to what that First the in air intensity (Da in
Cesium 137 is long enough that the
source may well outlive the process.
Another likelihood is that technologi-
cal advances will increase the sensitiv- 250
ity of the detector faster than the rate
at which the source is decaying. This 200
provides users the option of replacing
or upgrading the detector while keep-
150
ing the source in place for the future. Yr
Accumulated Rems
Unsafe s/
em
5R
• Radiation Safety 100 te:
Ra
The Nuclear Regulatory Commission
(NRC) limits radiation intensity to a 50 Safe
maximum of 5 milliroentgens per hour
(mr/hr) at a distance of 12 in from the 0
nuclear gage. If it is more, the area 10 20 30 40 50 60 70
Operator's Age
requires Radiation Area posting. The
distance of 12 in is critical, because radi-
Figure 9-8: Accumulation of Radiation Exposure
ation intensity decreases by the inverse
square of distance. Nuclear level gages person would receive if exposed to 2.4 mr/hr) is calculated at the detector,
are sized to provide radiation intensity mr/hr in the work area. As it is the total for the condition when there is no
at the detector that exceeds the mini- lifetime dosage of radiation exposure tank between the source and receiver.
mum required, but is under the 5 mr/hr that really matters (maximum of 250 Assume distance (d) is 48 in:
maximum. For ion chamber detectors, rems), the acceptability of the 5 rem/yr,
the minimum is 1 mr/hr. For Geiger- or any other limit, is also a function of Da = 1000 K(mCi)/d2 =
Mueller switches, it is 0.5 mr/hr. And age (Figure 9-8). On the other hand, the 1000(0.6)(10)/482 = 2.6 mr/hr
for scintillation detectors, it is 0.1-0.2 radiation at the detector must still be
mr/hr. Because the nuclear gage is sufficient to produce a usable change in Because the source is shielded in all
basically measuring the vapor space detector output when the level changes. directions except towards the tank,
TRANSACTIONS Volume 4 99
Radiation-Based Level Instrumentation 9
the operator who is working near the very much a function of the accurate trodes is applied. When the tube is
detector will receive the maximum knowledge of the installation details. exposed to gamma radiation, the gas
dosage when the tank is empty. The ionizes and the ionized particles carry
two H-in steel walls will reduce Da • Detector Options the current from one electrode to the
(% transmission of 1-in steel in Figure The simplest and oldest type of radia- other. The more gamma radiation
1 is 49%) to 0.49 x 2.6 = 1.27 mr/hr. tion detector is the Geiger-Muller reaches the gas in the tube, the more
This is below the allowable maxi- tube. This instrument is most often pulses are generated. The resulting
mum but above the minimum need- identified with the Geiger counters pulse rate is counted by the associated
ed by the detector. that make a loud and dramatic clicking electronic circuitry, which makes mea-
When the tank is full, the pres- sound when exposed to radiation. The surements in pulses per second.
ence of 30 in of water in the radia- working component of this detector is This detector can be used as a
tion path will reduce this maximum a metal cylinder that acts as one of the level switch if it is calibrated to
intensity to 0.045 mr/hr (0.035 x 1.9 electrodes and is filled with an inert engage or disengage a relay when
= 0.045). This reduction in intensity gas. A thin wire down the center acts radiation intensity indicates a high or
well exceeds the required 50% drop as the other electrode. Glass caps are low level condition. The G-M tube
needed for sensitive measurement.
Note that the source size could
have been cut in half if a Geiger-
Mueller detector were used. A scin-
Vessel
tillation detector would reduce Clips and
Vessel
source size 5- to 10-fold. Support
Clip
Plate
The source size can also be
reduced by locating the source in the Radiation
Path
tip of a probe inside the tank and
moving it relatively close to the wall. "U" Bolt
When large level ranges are to be
measured, a strip source can be used
instead of a point source. The accura- Source
and
Plan View
cy of most nuclear level gages is Holder Detector
around 1% of range. If accounting
accuracy is desired, the source and Max. Liquid Level
the detector can both be attached to
45 Max
motor driven tapes and positioned at Radiation
Path
the level (or at the interface level, if
the tank contains two liquids).
Fortunately, today’s computers can Min. Liquid Level
easily crunch the numbers and formu-
las of any combination of geometry
1/2"
and design criteria. The biggest chal- Platform
lenge is not the calculation, but the 30"
obtaining of accurate inputs for the
48"
calculations. Therefore, it is very
Elevation
important that your vessel’s wall mate-
rials, thicknesses, other tank compo-
Figure 9-9: Radiation Source Sizing
nents such as baffles, agitator blades
or jackets, and all distances be accu- used as insulators, and a high voltage detector can only be used as a single
rately determined. In short, the perfor- (700-1000 vdc) nearly sufficient to point detection device. Its advan-
mance of a nuclear gage installation is cause current flow between the elec- tages include its relatively low cost,
small size, and high reliability. as the intensity of gamma radiation measured by nuclear gages are among
The ion chamber detector is a con- increases. The photons travel through the most dangerous, highly pressur-
tinuous level device. It is a 4 to 6-in the clear plastic scintillator medium ized, toxic, corrosive, explosive, and
diameter tube up to twenty feet long to a photo multiplier tube, which carcinogenic materials around.
filled with inert gas pressurized to converts the light photons into elec- Because the nuclear gage “sees”
several atmospheres. A small bias trons. The output is directly propor- through tank walls, it can be installed
voltage is applied to a large elec- tional to the gamma energy that is and modified while the process is run-
trode inserted down the center of striking the scintillator. ning—without expensive down time
the ion chamber. As gamma energy Scintillators are available in a mul- or chance accidental release.
strikes the chamber, a very small sig- titude of shapes, sizes, and lengths. Because the installation of nuclear
nal (measured in picoamperes) is One of the latest is a fiber optic sensors requires a Nuclear
detected as the inert gas is ionized. cable that allows one to increase Regulatory Commission (NRC)
This current, which is proportional to detector sensitivity by installing license, associated procedures are
the amount of gamma radiation more filaments in the bundle. designed to guarantee that the
received by the detector, is amplified Another advantage of the fiber optic installation will be safe. The best way
and transmitted as the level mea- cable is that it is manufactured in to look at the safety aspects of
surement signal. long lengths flexible enough to radioactive gaging is to compare the
In level measurement applications, form-fit to the geometry of the ves- well defined and understood risk
the ion chamber will receive the most sel. This simplifies the measurement represented by exposing the opera-
radiation and, therefore, its output will of levels in spherical, conical, or tors to radiation against the possibly
be highest when the level is lowest. As other oddly shaped vessels. larger risk of having an unreliable or
the level rises and the greater quantity inaccurate level reading on a danger-
of measurand absorbs more gamma • Nuclear Applications ous process.
radiation, the output current of the Radiation gages typically are consid- As detectors become more sensi-
detector decreases proportionally. ered when nothing else will work, or tive and are aided by computers,
The system is calibrated to read 0% when process penetrations required radiation source sizes and the result-
level when the detector current out- by a traditional level sensor present a ing radiation levels continue to drop.
put is its highest. 100% level is set to risk to human life, to the environ- Therefore, the safety of these instru-
match the lowest value of the output ment, or could do major damage to ments is likely to continue to
current. Non-linearities in between property. The liquids and bulk solids improve with time. T
can usually be corrected with the use
of linearizing software. This software
can correct for the effects of steam References & Further Reading
coils, agitator blades, baffles, stiffening • OMEGA Complete Flow and Level Measurement Handbook and
rings, jackets and other components Encyclopedia®, OMEGA Press, 1995.
inside or outside the tank. • OMEGA Volume 29 Handbook & Encyclopedia, Purchasing Agents
Scintillation counter detectors are Edition, OMEGA Press, 1995.
five to ten times more sensitive than • Automated Process Control Electronics, John Harrington, Delmar
ion chambers. They also cost more, Publishing Inc., 1989.
yet many users are willing to accept • Fundamentals of Radar Techniques for Level Gauging, Detlef Brumbi,
the added expense because it allows Krohne Metechnik GmbH & Co. KG, 1995.
them either to use a smaller source • Industrial Applications of Radar Technology for Continuous Level
size or to obtain a more sensitive Measurement, W. L. Hendrick, Instrument Society of America, 1992.
gage. When gamma energy hits a scin- • Instrument Engineer’s Handbook, Bela Liptak, Third Edition, CRC Press, 1995.
tillator material (a phosphor), it is con- • Process/Industrial Instruments and Controls Handbook, 4th Edition,
verted into visible flashes comprised Douglas M. Considine, McGraw-Hill, 1993.
of light photons (particles of light). • Theoretical Nuclear Physics Volume I: Nuclear Structure, New York, A.
These photons increase in number deShalit, H. Feshback,: John Wiley & Sons, 1974.
One of the simplest thermal level typically 0.5 second and accuracy is suddenly increases. This is the level
switch designs consists of a temper- within 2 mm. In general, thermal level inside the mold. Using multiple sen-
ature sensor heated with a constant switches work best with non-coating sors spaced vertically, the system
amount of heat input. As long as the liquids and with slurries having 0.4-1.2 can determine the level of molten
probe is in the vapor space, the specific gravity and 1-300 cP viscosity. metal in the mold to within a frac-
probe remains at a high temperature, A third type of thermal switch tion of an inch.
because low-conductivity vapors do also uses two sensors inside the
not carry much heat away from the same vertical probe. One is mount- Vibrating Switches
probe. When the probe is sub- ed above the other and both are Vibrating level switches detect the
merged, the liquid absorbs more heat connected to a voltage source. dampening that occurs when a
and the probe temperature drops. When both are in the vapor or both vibrating probe is submerged in a
The switch is actuated when this in the liquid phase, the current flow process medium. The three types of
change in temperature occurs. vibrating sensors—reed, probe, and
Another type of thermal sensor tuning fork—are distinguished by
Pickup End Driver End
uses two resistance temperature their configurations and operating
detectors (RTDs), both mounted at frequencies (120, 200-400, and 85 Hz,
the same elevation. One probe is respectively). Their methods of oper-
heated and the other provides an ation and applications are similar.
unheated reference. The outputs of The reed switch consists of a paddle,
both sensors are fed into a a driver and a pickup (Figure 10-2).
Wheatstone bridge (Figure 10-1). The driver coil induces a 120-Hz
While the sensor is in the vapor Paddle Node vibration in the paddle that is
phase, the heated probe will be Figure 10-2: Vibrating Reed Switch damped out when the paddle gets
warmer than the reference probe, covered by a process material. The
and the bridge circuit will be unbal- through the two sensors is the same. switch can detect both rising and
anced. When both probes are sub- If, on the other hand, the lower one falling levels, and only its actuation
merged in the process liquid, their is in liquid and the upper in vapor, depth (the material depth over the
temperatures will approach that of the more current will flow through the paddle) increases as the density of
liquid. Their outputs will be nearly lower sensor. A current comparator the process fluid drops. The variation
equal and the bridge will be in balance. can detect this difference and signal in actuation depth is usually less than
This level switch is actuated when a that the sensor has reached the an inch. A reed switch can detect liq-
change in bridge balance occurs. vapor/liquid interface. uid/liquid, liquid/vapor, and
Since all process materials have a One interesting feature of this solid/vapor interfaces, and can also
characteristic heat transfer coeffi- design is that the sensor capsule can signal density or viscosity variations.
cient, thermal level switches can be be suspended by a cable into a tank or When used on wet powders, the
calibrated to detect the presence or well, and the sensor output can be vibrating paddle has a tendency to
absence of any fluid. Therefore, these used to drive the cable take-up motor. create a cavity in the granular solids.
switches can be used in difficult ser- In this fashion, the level switch can be If this occurs, false readings will
vices, such as interfaces, slurry, and used as a continuous detector of the result, because the sensor will con-
sludge applications. They can also location of the vapor/liquid interface. fuse the cavity with vapor space.
detect thermally conductive foams if Thermometers also can be used to It is best to use a reed switch on
spray-cleaned after each operation. detect level in higher temperature non-coating applications or to pro-
Thermal level and interface switches processes, such as measuring the vide automatic spray washing after
have no mechanical moving parts and level of molten steel in casting molds. each immersion in a sludge or slurry.
are rated for pressures up to 3,000 The thermometers do not actually Probe-type vibrating sensors are less
psig and process temperatures from touch the molten metal; instead, they sensitive to material build-up or coat-
-75 to 175°C (-100 to 350°F). When identify the place where the temper- ing. The vibrating probe is a round
detecting water level, response time is ature on the outside of the mold stainless steel element (resembling a
thermowell) that extends into the PVDF, polypropylene, stainless steel, non-contacting design.
material. If Teflon® coated and insert- carbon steel, and aluminum. They In a non-contacting, reflecting
ed at an angle, these devices tend to are available with Teflon® coatings or optical sensor, a beam of light is
be self-cleaning. Both the drive and in hygienic versions for sanitary aimed down at the surface of the
the sensor are piezoelectric ele- applications. process material. When the level of
ments: one causes the vibration and Vibrating sensors can be used to this surface rises to the setpoint of
the other measures it. When the ascertain liquid, solid, and slurry lev- the switch, the reflected light beam is
probe is buried under the process els. Reed switches can operate at detected by a photocell. Both the
LED light source and photodetector
are housed behind the same lens.
By adjusting the photocell or the
detection electronics, the sensor can
LED LED
Receiver be calibrated to detect levels at dis-
Receiver
Light
tances 0.25 to 12 in below the sensor.
from
Prism LED Prism These reflective switches can measure
Light
the levels of clear as well as translu-
Lost in
Liquid cent, reflective, and opaque liquids.
Some solids also can be detected. By
Liquid Below the
Liquid Immersing
using multiple photocells, a sensor
Sensing Prism the Sensing Prism can detect several levels.
Laser light also can be used when
Figure 10-3: Optical Level Switch making difficult level measurements,
material, its vibration is dampened pressures up to 3,000 psig, while tun- such as of molten metals, molten
and this decrease triggers the switch. ing forks and vibrating probes are glass, glass plate, or any other kind of
Vibrating probe sensors can be limited to 150 psig. Operating tem- solid or liquid material that has a
used to monitor powders, bulk peratures range from -100 to 150°C reflecting surface. If the receiver
solids, and granular materials such as (-150 to 300°F) and response time is module is motor driven, it can track
grain, flour, plastic pellets, cement, about 1 second. the reflected laser beam as the level
and fly ash. Their vibrating nature rises and falls, thereby acting as a
tends to minimize the bridging that Optical Switches continuous level transmitter.
occurs in solid materials. Tuning fork Using visible, infrared, or laser light, A refracting sensor relies on the
sensors are vibrated at about 85 Hz optical sensors rely upon the light principle that infrared or visible
by one piezoelectric crystal, while transmitting, reflecting, or refracting light changes direction (refracts)
another piezoelectric crystal detects properties of the process material when it passes through the interface
the vibration. As the process fluid when measuring its level. The optical between two media. When the sen-
rises to cover the tuning forks, the level switch can be of a contacting or sor is in the vapor phase, most of the
vibration frequency changes.
Like vibrating probes, tuning-fork
designs can be self-cleaning if
Teflon® coated and installed at an
angle. They can also be calibrated to
detect a wide range of materials,
including lubricating oils, hydraulic
fluids, water, corrosive materials,
sand, thick and turbulent fluids, pow-
ders, light granules, and pastes.
Tuning-fork sensors can be con-
structed with components made of Ultrasonic liquid level switches provide a 300:1 signal ratio from dry to wetted state.
light from the LED is reflected back encounter the sludge layer. peratures up to 125°C (260°F).
within a prism (Figure 10-3). When the Other transmission sensors rely on Response time is virtually immedi-
prism is submerged, most of the light the refraction principle utilizing an ate, and detection accuracy of most
refracts into the liquid, and the unclad, U-shaped fiber optic cable. A designs is within 1 mm. Optical level
amount of reflected light that reaches light source transmits a pulsed light switches are also designed for spe-
the receiver drops substantially. beam through the fiber cable, and cific or unique applications. For
Therefore, a drop in the reflected the sensor measures the amount of example, Teflon® optical level
light signal indicates contact with light that returns. If liquid covers the switches are available for sensing
the process liquid. cable, it will cause light to refract the level of ultra-pure fluids. Other
A refracting sensor cannot be used away from the cable. The use of fiber- unique designs include a level
with slurries or coating liquids, unless optics makes the system impervious switch that combines an optical
it is spray-washed after each submer- to electrical interference, and some with a conductivity-type level sen-
sion. Even a few drops of liquid on the designs are also intrinsically safe. sor to detect the presence of both
prism will refract light and cause erro- Optical sensors can operate at water (conductive) and hydrocar-
neous readings. Refracting sensors are pressures up to 500 psig and tem- bons (nonconductive). T
designed to be submerged in liquids;
therefore, any number of them can be
installed on a vertical pipe to detect a References & Further Reading
number of level points. • OMEGA Complete Flow and Level Measurement Handbook and
Transmission optical sensors send a Encyclopedia®, OMEGA Press, 1995.
beam of light across the tank. A • OMEGA Volume 29 Handbook & Encyclopedia, Purchasing Agents
sludge level sensor of this design uses Edition, OMEGA Press, 1995.
an LED and a photocell at the end of • Industrial Control Handbook, E.A. Parr, Butterworth-Heinemann Ltd., 1995.
a probe, located at the same eleva- • Instrument Engineer’s Handbook, Bela Liptak, Third Edition, CRC Press, 1995.
tion and separated by a few inches. To • Process/Industrial Instruments and Controls Handbook, 4th Edition,
find the sludge level, a mechanism (or Douglas M. Considine, McGraw-Hill, 1993.
an operator, manually) lowers the • The McGraw-Hill Encyclopedia of Science and Technology, 8th Edition,
probe into the tank until the sensors John H. Zifcak, McGraw-Hill, 1997.