AGFA CR30-X Service Manual
AGFA CR30-X Service Manual
AGFA CR30-X Service Manual
Service Manual
Imaging Services
DIGITIZERS
Type 5175 / 100/110
1st edition
CONFIDENTIALITY NOTE:
Use, dissemination, distribution or reproduction of this document by unauthorized personnel is not permitted and may be unlawful.
DOCUMENT CONTROL NOTE:
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
DD+DIS150.06E
Manufacturer
Agfa HealthCare N.V.
Tegernseer Landstrae 161
D - 81539 Mnchen
Germany
WARNING:
Improper operation or service activities may cause damage or injuries.
INSTRUCTION:
(1)
(2)
Strictly observe all safety directions within the "Generic Safety Directions" and on
the product.
Edition 1, Revision 2
10-2007
DIGITIZERS
Type 5175 / 100/110
Chapter 0 / 2
Agfa Company Confidential
DD+DIS150.06E
Document History
Edition.
Revision
Release
Date
Changes
compared to previous version 1.1
1.2
10-2007
Edition 1, Revision 2
10-2007
DIGITIZERS
Type 5175 / 100/110
Chapter 0 / 3
Agfa Company Confidential
DD+DIS150.06E
Chapter Overview
Chapter
0
Order List
Controls, Connections and Set Up Procedures
Functional Description
3.2
3.3
Troubleshooting
3.4
3.5
Replacement of Parts
3.6
3.7
3.8
3.9
Accessories
Field Modifications
Maintenance
10
Service Bulletins
11
Installation Planning
12
Glossary
Edition 1, Revision 2
10-2007
DIGITIZERS
Type 5175 / 100/110
Chapter 0 / 4
Agfa Company Confidential
DD+DIS150.06E
Explanation of notes
Signal Word
Situation
CAUTION:
WARNING:
DANGER:
Name
Type of Information
INSTRUCTION:
IMPORTANT:
NOTE:
Conventions
Actions
Way of writing
Sample
Action,
explanation
<omni-cd.exe>
Double-click the
<omni-cd.exe> icon
vips
Edition 1, Revision 2
10-2007
DIGITIZERS
Type 5175 / 100/110
Chapter 0 / 5
Agfa Company Confidential
DD+DIS150.06E
Subtype 100:
CR 30-X
Subtype 110:
To keep it simple this manual talks about "the digitizer" only. The digitizer CR 30-X is
always included, if not otherwise stated.
The digitizer is part of the digitizer system, consisting of .......
Digitizer
NX Workstation (Tower or laptop version)
Cassette with Image Plate (IP)
Digitizer
NX Workstation
Tower Version
NX Workstation
Laptop Version
Digitizer
Cassette
Cassette
517511ea.cdr
Figure 1
Edition 1, Revision 2
10-2007
DIGITIZERS
Type 5175 / 100/110
Chapter 0 / 6
Agfa Company Confidential
DD+DIS150.06E
NOTE:
This manual treats the digitizer with cassettes and image plates.
For more information to the other components refer to .....
Component
Information Location
NX Workstation
Printer
Edition 1, Revision 2
10-2007
DIGITIZERS
Type 5175 / 100/110
Chapter 0 / 7
Agfa Company Confidential
Published by
Agfa HealthCare N.V.
Tegernseer Landstrae 161
D - 81539 Mnchen
Germany
AGFA and the Agfa-Rhombus are trademarks of Agfa HealthCare N.V.
DOCUMENT CONTROL NOTE:
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
HEALTHCARE
Imaging Services
Document History
Edition.
Revision
Release
Date
Changes
compared to previous Revision 1.2:
1.3
07-2009
Referenced Documents
Document
Title
Not applicable
Not applicable
Edition 1, Revision 3
07-2009 printed in Germany
DD+DIS238.06E
Manufacturer
Agfa HealthCare N.V.
Publisher
Agfa-Gevaert HealthCare GmbH
Tegernseer Landstrae 161
D - 81539 Mnchen
Germany
Copyright 2009 Agfa HealthCare N.V.
All rights reserved.
Technical modifications reserved.
AGFA and the Agfa-Rhombus are trademarks of Agfa HealthCare N.V.
WARNING:
Improper operation or service activities may cause damage or injuries.
INSTRUCTION:
(1)
(2)
Strictly observe all safety directions within the "Generic Safety Directions" and on
the product.
Edition 1, Revision 3
07-2009
Page 2 of 28
Agfa Company Confidential
DD+DIS238.06E
LIST OF CONTENTS
DISCLAIMER......................................................................................................... 5
LABELS ................................................................................................................. 7
3.1
CE Mark................................................................................................... 7
3.2
3.3
REFERENCES .................................................................................................... 10
INTENDED USE.................................................................................................. 11
10
11
12
COMPLIANCE..................................................................................................... 14
13
14
15
16
17
18
19
20
Edition 1, Revision 3
07-2009
Page 3 of 28
Agfa Company Confidential
DD+DIS238.06E
21
22
23
24
RECYCLING ........................................................................................................27
25
26
Edition 1, Revision 3
07-2009
Page 4 of 28
Agfa Company Confidential
DD+DIS238.06E
Disclaimer
The installation and service of equipment described herein is to be performed by qualified
personnel who are employed by Agfa HealthCare or one of its affiliates or who are
otherwise authorized by Agfa HealthCare or one of its affiliates to provide such services.
Fitters, engineers and other persons who are not employed by or otherwise directly
affiliated with or authorized by Agfa HealthCare or one of its affiliates are directed to
contact one of the local offices of Agfa HealthCare or one of its affiliates before attempting
installation or service procedures.
No part of this document may be reproduced, copied, adapted or transmitted in any form
or by any means without the written permission of Agfa HealthCare.
Agfa HealthCare makes no warranties or representation, expressed or implied, with
respect to the accuracy, completeness or usefulness of the information contained in this
document and specifically disclaims warranties of suitability for any particular purpose.
Agfa HealthCare shall under no circumstances be liable for any damage arising from the
use or inability to use any information, apparatus, method or process disclosed in this
document.
Agfa HealthCare is not liable for resulting consequences, damages or injuries if you dont
operate the product correctly or if you dont have it serviced correctly.
Agfa HealthCare reserves the right to change the product, the characteristics and its
documentation without further notice to improve reliability, function or design.
NOTE:
In the United States, Federal Law stipulates that medical devices should only be sold to,
distributed and used by or by order of a licensed physician.
Edition 1, Revision 3
07-2009
Page 5 of 28
Agfa Company Confidential
DD+DIS238.06E
Used Icons
Icon
Edition 1, Revision 3
07-2009
Page 6 of 28
Agfa Company Confidential
DD+DIS238.06E
Labels
3.1
CE Mark
CE Mark
3.2
System Labels
All system labels and software version number locations are referred to within this service
document in the appropriate section.
Enclosed an overview of common labels, according to ISO 3864.
This list is not complete.
Hot Surface
Obstacles
Laser Beam
Corrosive Liquid
High Voltage
Hand Injuries
C&W_005.cdr
Edition 1, Revision 3
07-2009
Page 7 of 28
Agfa Company Confidential
DD+DIS238.06E
3.3
Class #
Meaning
Class 1:
Class 1 M:
Class 2:
Example Label
LASER RADIATION
DO NOT VIEW DIRECTLY WITH
OPTICAL INSTRUMENTS
CLASS 1M LASER PRODUCT
LASER RADIATION
DO NOT STARE INTO BEAM
CLASS 2 LASER PRODUCT
Edition 1, Revision 3
07-2009
Page 8 of 28
Agfa Company Confidential
DD+DIS238.06E
Class #
Meaning
Class 2M:
Class 3R:
Example Label
LASER RADIATION
DO NOT STARE INTO THE BEAM
OR VIEW DIRECTLY WITH
OPTICAL INSTRUMENTS
CLASS 2M LASER PRODUCT
LASER RADIATION
AVOID DIRECT EYE EXPOSURE
CLASS 3R LASER PRODUCT
Class 3B:
LASER RADIATION
AVOID EXPOSURE TO BEAM
CLASS 3B LASER PRODUCT
Class 4:
Edition 1, Revision 3
07-2009
Page 9 of 28
Agfa Company Confidential
DD+DIS238.06E
Product Complaints
Any service person who has any complaints or has experienced any dissatisfaction in the
quality, durability, reliability, safety, effectiveness or performance of this product must
notify Agfa HealthCare by the Agfa HealthCare complaint procedure.
If the product malfunctions and may have caused or contributed to a serious injury of a
patient or an accident or if there are any hazards which may cause an accident
Agfa HealthCare must be notified immediately by telephone, fax or written
correspondence to the following address:
Agfa Service Support - local support addresses and phone numbers listed on:
www.agfa.com
References
Technical Documentation is available via MedNet (PDF) and your local Agfa HealthCare
support organisation (Paper).
Access to MedNet:
IntraNet: http://docs.agfanet/bu/mi/mednet/mednetcso.nsf
ExtraNet: http://extranet.agfa.com/bu/mi/mednet/mednetcso.nsf
Edition 1, Revision 3
07-2009
Page 10 of 28
Agfa Company Confidential
DD+DIS238.06E
Intended Use
This Agfa HealthCare product should only be operated in a hospital or clinical radiological
environment by qualified staff.
It must only be operated according to its specifications and its intended use. Any
operation not corresponding to the specifications or intended use may result in hazards,
which in turn may lead to serious injuries or fatal accidents (for example electric shocks).
AGFA will not assume any liability whatsoever in these cases.
Make sure that the product is constantly monitored in order to avoid inappropriate
handling, especially by children.
The product must only be installed and put into operation under the specified conditions.
Intended User
This manual is written for Agfa trained Field Service Engineers and Clinical Application
Specialists, trained users of Agfa HealthCare products and trained diagnostic XRay
clinical personnel who have received proper training. Users are considered as the persons
who handle the equipment as well as the persons having authority over the equipment.
Edition 1, Revision 3
07-2009
Page 11 of 28
Agfa Company Confidential
DD+DIS238.06E
Must make his or her personal contribution to improve safety and protect the
environment.
When working on a customers site, has a duty to take reasonable care to avoid injury
to himself or herself or to others who may be affected by their acts or omissions.
Is obligated to adhere strictly to regulations and instructions.
Shall familiarise himself or herself with the provisions of the Agfa Healthcare
Health, Safety and Environment Policy and any specific rules or procedures relating to
occupational safety at work and the protection of the environment.
Shall promptly report any near misses, accidents, incidents or dangerous occurrences
to their line manager and co-operate fully in any investigation.
Shall co-operate with company management on matters relating to
health, safety and environment and, where appropriate, discuss with and / or assist
their manager in resolving matters relating to health, safety and environment.
Shall ensure that any company equipment issued to them, or, for which they are
responsible, is correctly used and properly maintained.
Shall wear protective equipment whenever instructed or if it is recommended to do so.
Shall be responsible for good housekeeping in the area in which he or she is working.
Shall report situations, which could put them at risk, on either company or
customers' premises, to their manager or supervisor; and, if warranted, directly and in
confidence, to the Health and Safety Co-ordinator, Global HSE Manager, or ultimately
to the Managing Director.
Shall report any injuries, diseases or dangerous occurrences to his or her line
manager.
Shall report any accidents, incidents or near misses to his or her line manager.
Shall report any situation of which he or she is aware that is potentially dangerous.
Shall comply with any health surveillance procedure instituted for his or her benefit or
for compliance with regulations.
Edition 1, Revision 3
07-2009
Page 12 of 28
Agfa Company Confidential
DD+DIS238.06E
10
INSTRUCTION:
Consult the Technical Documentation before making any connections to other equipment.
Consideration relating to the choice of accessory equipment shall include:
In addition all configurations must comply with the medical electrical systems standard IEC
60601-1-1. The party that makes the connections acts as system Configurer and is
responsible for complying with the systems standard.
If required, contact your local service organization.
Edition 1, Revision 3
07-2009
Page 13 of 28
Agfa Company Confidential
DD+DIS238.06E
11
INSTRUCTION:
12
Use only tools and measuring instruments which are suitable for the procedure.
Only approved Agfa HealthCare accessories must be used. For a list of compatible
accessories contact your local Agfa HealthCare organization or www.agfa.com.
Compliance
Directive for HealthCare Imaging Products:
Council Directive 93/42/EEC of 14 June 1993 concerning medical devices
(OJ No L 169/1 of 1993-07-12)
Edition 1, Revision 3
07-2009
Page 14 of 28
Agfa Company Confidential
DD+DIS238.06E
Edition 1, Revision 3
07-2009
Page 15 of 28
Agfa Company Confidential
DD+DIS238.06E
Note:
This product has been tested and found to comply with the limits for a Class A
computing device pursuant to Subpart B of Part 15 of FCC Rules, which are designed to
provide reasonable protection against such interference when operated in a commercial
environment.
Operation of this equipment in a residential area is likely to cause interference.
The user will be required to take all necessary measures to correct the interference at
his own expense.
Edition 1, Revision 3
07-2009
Page 16 of 28
Agfa Company Confidential
DD+DIS238.06E
13
Under certain conditions the Agfa HealthCare product will show a display containing a
message. This message will show that either a problem or action has occurred or that
a requested action is required or cannot be performed. The user must read these
messages carefully they will provide information on what to do. This will be either
performing an action to resolve the problem or to contact the Agfa HealthCare
service organization. Details on the contents of messages can be found in this
Technical Documentation.
All images created using any image technology can show artifacts which could be
confused with diagnostic information. If there is any doubt that the diagnostic
information could be corrupted, additional investigations must be performed to get
clear diagnostic information.
Ventilation openings must not be covered.
If you notice conspicuous noise or smoke, disconnect the product immediately from
the mains.
Do not pour water or any other liquid over the device.
If a system malfunction causes an emergency situation involving the patient, operating
personnel or any system component, activate the emergency stop for the system
concerned. All motor driven system movements will be stopped.
Do not store any magnetic media near or on devices, which produce magnetic fields,
since stored data may be lost.
Explosive environment:
DANGER:
Risk of explosion.
Never operate this device in zones where there are flammable anesthetics or oxygen which
may cause an explosion.
Edition 1, Revision 3
07-2009
Page 17 of 28
Agfa Company Confidential
DD+DIS238.06E
14
Radiation Protection
Only qualified and authorized personnel shall operate any X-Ray system. In this context
qualified means those legally permitted to operate this X-Ray equipment in the jurisdiction
in which the X-Ray equipment is being used, and authorized means those authorized by
the authority controlling the use of the X-Ray equipment. Full use must be made of all
radiation protection features, devices, systems, procedures and accessories.
Ionizing radiation can lead to radiation injuries if handled incorrectly. When radiation is
applied, the required protective measures must be complied with.
15
Details about cleaning and disinfection or sterilization methods that may be used on
SYSTEM parts or ACCESSORIES that can become contaminated through contact
with the PATIENT or with body fluids, are referred to within the individual service
documents.
Disconnect the power supply from the equipment prior to cleaning the equipment.
Edition 1, Revision 3
07-2009
Page 18 of 28
Agfa Company Confidential
DD+DIS238.06E
16
This system uses high voltage. Please consider the respective safety regulations.
Electrical repairs and connections must only be performed by a qualified electrician.
Mechanical repairs and connections must only be performed by a qualified technician.
The safety directions for operation (see section 13) are also valid for all service
activities.
During all service activities observe prescribed local and country-specific
requirements (e.g. occupational safety and accident prevention regulations).
All existing screw connections must be tightened sufficiently firmly, but they may not
be overstressed when tightening. There must always be compliance with stated
torque values!
Damaged or missing screws may be replaced only with the same screw types that
have the specified hardness rating. Unless a different value is listed in the
instructions, all Allen screws used must be hardness rated 8.8.
All screws must be secured in accordance with the corresponding data.
If "Loctite" has to be used to secure screws, this is stated in the text.
Any Agfa service PC or tool which is to be connected via RS232, RJ45, USB or other
interface to an Agfa device must not be connected to the mains but must be operated
on its internal battery or indirect supply (low voltage).
When handling printed circuit boards (abbr.: PCBs) the following points must be
observed:
o Always switch off the equipment and unplug the power cord, before you
disconnect or connect cables on printed circuit boards.
o When working on PCBs, always wear an anti-static wrist strap. Never touch any
parts or components on PCBs with your bare fingers.
o PCBs have to be kept or transported in their protection bags. Never carry a
PCB without protection bag and walk on carpet or plastic floor covering
(electrostatic charge).
o Once the PCB is taken out of its protection bag, it has to be protected from
electrostatic charge by a grounded mat.
Edition 1, Revision 3
07-2009
Page 19 of 28
Agfa Company Confidential
DD+DIS238.06E
17
The digitizer and the cassette storage shall be protected against X-ray radiation this way,
that the annual dose equivalent at the installation place will not exceed 1 mSv.
The film-screen system shall be protected against X-ray radiation this way, that the annual
dose equivalent at the installation place will not exceed 1 mSv.
Local and International wiring regulations must be observed. Check all supplies and
voltages, currents, trips and fuses with the Hospital facilities department or their
engineers.
Edition 1, Revision 3
07-2009
Page 20 of 28
Agfa Company Confidential
DD+DIS238.06E
The device complies with the EN 60601-1, 2006 standard for Information Technology.
This means that, although it is absolutely safe, patients may not come in direct contact
with the equipment. Therefore the operator console must be placed outside a radius
of 1.5 m around the patient.
R = 1.5m
1.5m
1.5m
Patient environment
This device should be installed behind the institution firewall for network security and
anti-virus protection. No ongoing computer virus protection or network security for this
medical device is provided (e.g. a computer firewall). Network security and anti-virus
provisions are the ongoing responsibility of the user or institution.
Edition 1, Revision 3
07-2009
Page 21 of 28
Agfa Company Confidential
DD+DIS238.06E
18
Compare the power requirements indicated on the type label with the available
power supply in the installation room.
Check the service manual for the type of input voltage selection, manual or
automatic: If manual, select the appropriate voltage and fuses.
Confirm to use the correct socket and plug for the required power supply.
Check the equipment will work with the power supply available.
Edition 1, Revision 3
07-2009
Page 22 of 28
Agfa Company Confidential
DD+DIS238.06E
19
Always connect the associated monitor to the same Uninterruptible Power Source as
the PC.
This Technical Documentation identifies the parts on which preventive inspection and
maintenance shall be performed by Agfa HealthCare service personnel, including the
periods to be applied.
In general the device has to be switched off during service activities. Exception: If the
device is switched on to perform tests pay particular attention to any hazards due to
moving and rotating parts. Avoid lose clothing or finger traps. Switch off the device
immediately after the tests.
Do not turn motors manually. If required, first disconnect the motor from the motor
control board.
Make sure that the power cord does not show any signs of damage.
After repair work always check that the integrated safety features are not overridden
or disconnected.
If there is any visible damage to the machine casing do not hand-over the product to
the customer. First repair the machine casing.
Edition 1, Revision 3
07-2009
Page 23 of 28
Agfa Company Confidential
DD+DIS238.06E
Replacing batteries:
WARNING:
Battery can explode, causing chemical burns.
INSTRUCTION:
Only use batteries of the same type or an equivalent type as specified by the
manufacturer.
When removing lithium batteries from the equipment take appropriate measures to
avoid short circuit of the battery:
Either use tape to cover the two poles of the battery or put the battery back in its
original packing and secure the packing by tape.
After any work at the power supply or at any component connected to mains voltage
inform the responsible organization 1 about the necessity of the electrical test
according to national regulations.
If specific national regulations do not exist: It is recommended to perform the
electrical test according to IEC 62353.
Make sure, that all grounding connections to metallic covers and all grounding
connections inside the device are present.
NOTE:
On MedNet, GSO Library path "General Info => Agfa HealthCare => Publications =>
Service Manual" a form IEC 62353:2007 Test Documentation for CR Digitizers is
available, to be handed out to the responsible organization (Intranet Link / Extranet Link).
Responsible Organization:
Entity accountable for the use and maintenance of a medical equipment or a medical equipment system. The
accountable entity can be, for example, a hospital or an individual clinician.
DOCUMENT CONTROL NOTE:
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 3
07-2009
Page 24 of 28
Agfa Company Confidential
DD+DIS238.06E
Strictly observe the warning notes in the service manual of devices emitting laser
radiation (See service manual chapter describing Safety Guidelines / General
Repair Instructions) and at the corresponding steps of instructions.
Strictly observe the warning labels at the modules emitting laser light. For the
meaning of the labels refer to section 3.3 in this document.
Do not open modules containing a laser. Only open modules containing a laser if
explicitly instructed to do so.
Do not keep tools in the laser beam unless explicitly instructed to do so.
Make yourself familiar with the path of the laser light and the conditions, when the
laser beam is switched on. Refer to the Functional Description in the
corresponding service manual.
Sharp edges:
CAUTION:
Sharp edges inside the device: Cut or abrasion possible.
Be careful at maintenance and replacement of parts.
Edition 1, Revision 3
07-2009
Page 25 of 28
Agfa Company Confidential
DD+DIS238.06E
Secured screws:
CAUTION:
Opening screws secured by red lacquer may misalign important device
adjustments:
Do not open screws that are secured by red lacquer.
Replacing fuses:
Warning:
Replacing fuses by wrong type may lead to fire hazard!
Use only fuses of the exact value and characteristics stated in the service manual or on
the device.
20
Edition 1, Revision 3
07-2009
Page 26 of 28
Agfa Company Confidential
DD+DIS238.06E
21
22
23
24
Recycling
Agfa HealthCare has Recycling Passports available for all equipment. The Recycling
Passport explains whether hazardous materials, special components and batteries are
present, where they are located and how they can be removed at the end of the life cycle.
The Recycling Passports are meant to be used as information for waste treatment
partners and companies that want to recycle end-of-life Agfa equipment.
To get a copy of the required Agfa HealthCare Recycling Passport please contact your
local Sales organization.
Edition 1, Revision 3
07-2009
Page 27 of 28
Agfa Company Confidential
DD+DIS238.06E
25
Waste Disposal
On August 13, 2005, the European Directive on Waste Electrical and
Electronic Equipment (WEEE) 2002/96/EC, amended by Directive
2003/108/EC, came into force.
The directive on Waste Electrical and Electronic Equipment (WEEE) aims to
prevent the generation of electric and electronic waste and to promote the
reuse, recycling and other forms of recovery. It therefore requires the collection
of WEEE, recovery and reuse or recycling.
This directive has to be implemented into national law by the individual
European countries by August 13th 2005.
Due to the implementation into national law, specific requirements can be
different within the European Member States.
This symbol on the product, or in the manual and in the warranty, and / or on
its packaging indicates that this product shall not be treated as household
waste.
For more detailed information about take-back and recycling of this product,
please contact your local Agfa service organization. By ensuring this product is
disposed of correctly, you will help prevent potential negative consequences
for the environment and human health, which could otherwise be caused by
inappropriate waste handling of this product. The recycling of materials will
help to conserve natural resources.
If your equipment or replaced spare parts contain batteries or accumulators
please dispose of these separately according to local regulations.
26
Edition 1, Revision 3
07-2009
Page 28 of 28
Agfa Company Confidential
Chapter 1
HEALTHCARE
Controls, Connections,
and Setup Procedures
Imaging Services
CR 30-X
Type 5175 / 100/110
all important routines to be carried out prior to putting the machine in operation
Document History
Edition.
Revision
Release
Date
Changes
compared to previous Version 1.3:
1.4
07-2008
Referenced Documents
Document
Title
n.a.
n.a.
Edition 1, Revision 4
07-2008 printed in Germany
DD+DIS150.06E
Manufacturer
Agfa HealthCare N.V.
Publisher
Agfa-Gevaert HealthCare GmbH
Tegernseer Landstrae 161
D - 81539 Mnchen
Germany
Copyright 2008 Agfa HealthCare N.V.
All rights reserved.
Technical modifications reserved.
AGFA and the Agfa-Rhombus are trademarks of Agfa HealthCare N.V.
WARNING:
Improper operation or service activities may cause damage or injuries.
INSTRUCTION:
(1)
(2)
Strictly observe all safety directions within the "Generic Safety Directions" and on
the product.
NOTE:
To verify the latest version of single documents and of Service Manuals refer to the
Document Type Order List in the GSO library.
Edition 1, Revision 4
07-2008
CR 30-X
Type 5175 / 100/110
Chapter 1 / 2
DD+DIS150.06E
LIST OF CONTENTS
1.2
1.3
1.4
1.5
Mounting the Distance Holder with FireWire Cable Strain Relief .............................................8
1.6
1.7
1.7.1
1.7.2
INSTALLATION ......................................................................................................................12
2.1
2.2
NX Workstation Setup.............................................................................................................16
2.3
2.4
3.2
3.2.1
3.2.2
3.2.3
3.3
3.4
3.5
3.6
Filling in "Site and System Data" and "Installation Report" Forms .........................................25
Edition 1, Revision 4
07-2008
CR 30-X
Type 5175 / 100/110
Chapter 1 / 3
DD+DIS150.06E
NOTE:
The Installation Checklist in section 4 of this chapter gives an overview of all steps and
guides through the complete installation process.
Installation Prerequisites
The purpose of this section is to show the required tasks before installation to
guarantee a smooth digitizer installation.
The following table lists these tasks:
Task
Details see
1.1
1.2
1.3
1.4
1.5
1.6
1.7
1.1
1.2
Digitizer Unpacking
(1) Compare the labels on the boxes with the customer's order list and the shipping
papers.
(2) Check the packing material for visible transport damage such as
- dented edges
- damage on the box
- torn fixing elements (metal straps).
DOCUMENT CONTROL NOTE:
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 4
07-2008
CR 30-X
Type 5175 / 100/110
Chapter 1 / 4
DD+DIS150.06E
517501aa.cdr
Figure 1
IMPORTANT:
If the machine was tilted, the circle in the arrow head of the TILTWATCH changes from
white to red.
If the machine was subjected to shocks, the square field in the middle of the
SHOCKWATCH changes from white to red.
If damage is visible from the outside contact your local AGFA representative.
(4)
Edition 1, Revision 4
07-2008
CR 30-X
Type 5175 / 100/110
Chapter 1 / 5
DD+DIS150.06E
Note
Hand protection
Foot protection
F7.5175.0048.0
No. only needed for production purpose
1 piece 18TO4 MA 1
* 1 1 8 TO4 1 *
DD+DIS136.06E
SN
Type
ABC-Code
Device
3x
100 kg
Figure 2
Edition 1, Revision 4
07-2008
CR 30-X
Type 5175 / 100/110
Chapter 1 / 6
DD+DIS150.06E
1.3
1.4
2.
In case it is not complete inform the national service manager. He will decide how
to proceed.
WARNING:
Excessive vibrations during scanning may decrease image quality.
The structure and stability of the used table need to be suitable in relation with the
size and weight of the system. The table may not be subject to excessive shock and
vibrations from other sources.
For more information refer to the Installation Planning document DD+DIS172.06E.
(1)
(2)
If the optional Agfa table with wheels is used: Lock the wheels.
NOTE:
The transportation lock (= cable tie to fix
photomultiplier with light collector at PMI
carrier) does not have to be removed.
Transportation Lock
517511ah.cdr
Figure 3
Edition 1, Revision 4
07-2008
CR 30-X
Type 5175 / 100/110
Chapter 1 / 7
DD+DIS150.06E
1.5
2 screws M4 x 8 for
mounting distance holder
517501ww.cdr
Figure 4
(1)
517501wy.cdr
Figure 5
(2)
517501wz.cdr
Figure 6
Edition 1, Revision 4
07-2008
CR 30-X
Type 5175 / 100/110
Chapter 1 / 8
DD+DIS150.06E
1.6
(1) Check whether the correct UPS is shipped for the site:
Status 07/2008 for type of the optional UPS:
ABC Code
UPS Type
Rated Voltage
EGPSE
120 = Default
230 = Default
EGPTG
IMPORTANT:
When a different UPS is used, check whether it fits to the power requirements of the
connected devices.
(2) Unpack the UPS.
(3) Make the cabling as shown in Figure 7 and Figure 8 respectively.
517511aj.cdr
to wall outlet
517501af.cdr
Figure 7
DOCUMENT CONTROL NOTE:
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 4
07-2008
CR 30-X
Type 5175 / 100/110
Chapter 1 / 9
DD+DIS150.06E
USB
to wall outlet
517501ae.cdr
Figure 8
(4) Adapt dip switches to available output voltage. Refer to the UPS user's guide
which is part of delivery.
(5) On the control PC install LAN safe SW and USB drivers according to the
instructions provided with the Powerware software suite CD.
1.7
WARNING:
Using different power circles may lead to ground potential differences: In this
case the leakage current via FireWire cable may exceed the limits as defined
by IEC 60601:
Confirm that both, control PC and digitizer are connected to the same ground, e.g.
via multiple socket or UPS.
When connecting digitizer and control PC at two different sockets, installed in the
wall: Inform the hospital technician that digitizer and control PC have to be
connected to the same ground and the same phase.
DOCUMENT CONTROL NOTE:
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 4
07-2008
CR 30-X
Type 5175 / 100/110
Chapter 1 / 10
DD+DIS150.06E
1.7.1
Figure 9
1.7.2
NX Workstation
Laptop Version
Digitizer
6 pin to 4 pin
converter
517511ak.cdr
Figure 10
DOCUMENT CONTROL NOTE:
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 4
07-2008
CR 30-X
Type 5175 / 100/110
Chapter 1 / 11
DD+DIS150.06E
(4)
Figure 11
IMPORTANT:
Do not place the laptop on top of the digitizer, because the accumulated heat
developed by the digitizer and the laptop may affect the operation of the system.
Installation
The purpose of this section is to show the required installation tasks.
The following table lists these tasks:
Task
Details see
2.1
2.2
2.3
2.4
NOTE
The PC to which the digitizer is connected to is further named in this document:
Edition 1, Revision 4
07-2008
CR 30-X
Type 5175 / 100/110
Chapter 1 / 12
DD+DIS150.06E
2.1
(2)
517501ab.cdr
Figure 12
NOTE:
After switch on a selftest is executed.
If the NX workstation is operational it takes
approx. 1 minute until the status indicator
at the front panel changes from red to
green.
See Table 1 for meaning of the status
indicator colors.
Figure 13
Status Indicator
Blue - Constant
Green - Constant
Digitizer Status
Erasure cycle active
Ready
Green - Flashing
Red - Constant
Red - Flashing
Edition 1, Revision 4
07-2008
CR 30-X
Type 5175 / 100/110
Chapter 1 / 13
DD+DIS150.06E
IMPORTANT:
As the serial number of the digitizer is not yet known, approx. 30 secs. after first
switch on error 4637 (Serial Number not registered 1) and possibly error 2712 (ASAP
Software versions loaded onto machine and installed on PC are not identical) will be
displayed in the digitizer error viewer on the control PC.
Error 4637 will be solved by the registration of the digitizer serial number on
the control PC as described in this section.
Error 2712 (if present) will be solved with a check of the installed software
version, see step 2.4 Checking Software Version.
(Error 2712)
517511aj.cdr
Figure 14
(3)
(4)
(5)
(6)
(7)
Select OK.
1 In software < CRE_1509 error 4637 was described as Registration of Serial Number failed in the Explain Error Code
window of the CR 30-X service client.
DOCUMENT CONTROL NOTE:
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 4
07-2008
CR 30-X
Type 5175 / 100/110
Chapter 1 / 14
DD+DIS150.06E
Figure 15
Edition 1, Revision 4
07-2008
CR 30-X
Type 5175 / 100/110
Chapter 1 / 15
DD+DIS150.06E
2.2
NX Workstation Setup
NOTE
The NX workstation comes with a default CR 30-X System Configuration and
Exam-Tree XML File installed and activated.
The steps outlined here are only to confirm that the NX workstation was prepared by
production for the CR 30-X system.
(1) Switch on the NX workstation
(2) Log in as "crservice"
(3) In the main menu select "Service and Configuration Tools"
(4) Select Load active configuration
(5) Check menu Device Configuration Digitizer:
The CR 30-X must be available.
See also important note below.
(6) Continue with setup of the NX workstation as described in the appendix,
DD+DIS326.06E: Guideline for Set-Up and Customer Training
IMPORTANT:
The default DIGITIZER system configuration file has following settings:
Device Configuration:
o Digitizer CR 30-X incl. Digitizer Emergency button configured
o Printer Drystar 5302 configured
o Sheet and image text box configured (and alternatively one factory
default textbox)
Configuration of Examination:
o Examtree English + adapted Add image pane
o System diagnosis examination configured
(visible in Add image pane)
o Dedicated examination for the digitizer emergency button, so that the
examtree can be deleted without loosing the configuration of the digitizer
emergency button
o User interface pre-configured. Note: With import of an other examtree
the "add image pane" has to be adapted
The digitizer gets a default CR 30-X Configuration and Exam-Tree XML File loaded
in production.
A copy of this file is also available on MEDNET, GSO Library Computed
Radiography CR Digitizers CR 30-X Freeware
Edition 1, Revision 4
07-2008
CR 30-X
Type 5175 / 100/110
Chapter 1 / 16
DD+DIS150.06E
2.3
(2)
(3)
(4)
(5)
Figure 16
2.4
(1)
(2)
Edition 1, Revision 4
07-2008
CR 30-X
Type 5175 / 100/110
Chapter 1 / 17
DD+DIS150.06E
Node
PMI Board
HDL Board
SSC Board
Status
different
different
different
Select All
Clear Selection
517501da.cdr
Digitizer
NX Workstation
Figure 17
Result:
Edition 1, Revision 4
07-2008
CR 30-X
Type 5175 / 100/110
Chapter 1 / 18
DD+DIS150.06E
Completion of Installation
The purpose of this section is to show the tasks that are required to prepare the
digitizer for hand-over to the clinical application specialist.
The following table lists these tasks.
Task
3.1
Details see
3.1
3.2
3.3
3.4
3.5
3.6
* The last digit in the spare part number indicates the spare part revision at release of
this document. When ordering, the actual revision of the spare part is delivered.
Edition 1, Revision 4
07-2008
CR 30-X
Type 5175 / 100/110
Chapter 1 / 19
DD+DIS150.06E
(1)
517509ag.cdr
Figure 18
(2)
(3)
(4)
1
2
517509ad.cdr
Figure 19
(5)
(6)
(7)
(8)
517509ae.cdr
Figure 20
Moisten the cloth with cleaner and wipe the image plate surface softly and
evenly (phosphorous side and rear side).
Leave the image plate surface to dry for about 10 minutes to allow the
solvents to evaporate.
Use only a dry cloth or blow away possible dust with canned air to clean the
tray thoroughly. DO NOT USE A SCREEN CLEANER TO CLEAN THE
TRAY.
Once the image plate surface is dry (after about 10 minutes), check once
again for particles of material and other impurities..
Edition 1, Revision 4
07-2008
CR 30-X
Type 5175 / 100/110
Chapter 1 / 20
DD+DIS150.06E
(9)
3.2
Figure 21
3.2.1
Step
Details see
3.2.1
3.2.2
3.2.3
X-Ray tube
Rotating
Anode
Cathode
517509aa.cdr
NOT OK
OK
Figure 22
Edition 1, Revision 4
07-2008
CR 30-X
Type 5175 / 100/110
Chapter 1 / 21
DD+DIS150.06E
12 mAs
75 kVp
large focus.
The values are approximate and may vary among the X-ray devices.
(4)
(5)
(6)
(7)
IMPORTANT:
By selection of examination type "System Diagnosis - Flatfield", the correct settings
for image processing of the flatfield are used. Any other examination type may give a
result which cannot be compared with the limit pattern.
3.2.2
Edition 1, Revision 4
07-2008
CR 30-X
Type 5175 / 100/110
Chapter 1 / 22
DD+DIS150.06E
3.2.3
Figure 23
Evaluation:
If the scanned image visually looks as good or better than the limit pattern,
the image quality is okay.
If there are unacceptable effects in slow scan or fast scan direction see
chapter 3.3, Troubleshooting.
NOTE:
The slow scan direction is always the long side of the image plate.
30 cm
43 cm
30 cm
15 cm
24 cm
Slow scan 35 cm
direction
24 cm
18 cm
517509ab.cdr
Edition 1, Revision 4
07-2008
CR 30-X
Type 5175 / 100/110
Chapter 1 / 23
DD+DIS150.06E
IMPORTANT:
If you use a FLFS cassette for flatfield exposure, the image will be slightly darker in
the upper and lower image area where the backscatter protection is removed (approx.
1 cm each). These darker zones also have to be used for image quality evaluation in
slow scan and fast scan direction. See next page.
Flatfield with Genrad Cassette
Effect exaggerated
Genrad Cassette
FLFS Cassette
Backscatter
Protection
FLFS
10 mm
10 mm
FLFS
Figure 25
Edition 1, Revision 4
07-2008
CR 30-X
Type 5175 / 100/110
Chapter 1 / 24
DD+DIS150.06E
3.3
3.4
3.5
(1)
(2)
(3)
(4)
Save HOT file in directory "C:/Documents and Settings/ All Users /Application
Data/CR 30-X/log/" as "HOT_ddmmyyyy.txt".
3.6
Edition 1, Revision 4
07-2008
CR 30-X
Type 5175 / 100/110
Chapter 1 / 25
DD+DIS150.06E
Installation Checklist
#
Step
refer to ....
Okay
Installation Prerequisites:
1.
1.1
2.
Unpack digitizer.
1.2
3.
1.3
4.
1.4
5.
1.5
6.
1.6
7.
1.7
Installation
8.
2.1
9.
2.2
2.3
2.4
Completion of installation
12. Clean image plates.
3.1
3.2
3.3
3.4
3.5
17. Fill in "Site and System Data" and "Installation Report" form.
3.6
Edition 1, Revision 4
07-2008
CR 30-X
Type 5175 / 100/110
Chapter 1 / 26
DD+DIS150.06E
Functional Description
Chapter 2
List of Contents
1.1
1.2
CR30-X in a Nutshell....................................................................................... 1
1.3
NX Workstation in a Nutshell......................................................................... 2
1.4
Scanning .......................................................................................7
5.1
5.2
5.3
Edition 1, Revision 0
CR30-X
Type 5175/100
Chapter 2 / I
Agfa Company Confidential
DD+DIS150.06E
Functional Description
1.1
System Overview
The CR30-X digitizer is part of the
"CR 30-X System" which
comprises:
Digitizer CR30-X
NX Workstation
Cassette with Image Plate
(IP)
Digitizer
Cassette
1.2
NX Workstation
517511ae.cdr
Figure 1
CR30-X in a Nutshell
Photo Multiplier Interface
(PMI) Board
Optic Module
Erasure Unit
Cassette
Unit
Polygon Mirror
Light
Collector
Laser Diode
Polygon Motor
Chip
Reader
Laser
Beam
Slow Scan
Unit
IP Run
517511ag.cdr
Transport
Unit
FireWire
Cable to
NX Workstation
Calibration
Board
Edition 1, Revision 0
CR30-X
Type 5175/100
Chapter 2 / 1
Agfa Company Confidential
DD+DIS150.06E
Functional Description
1.3
NX Workstation in a Nutshell
The NX Workstation is a WIN2003 Server or WIN XP based Processing
Station with the NX software installed.
Every incoming image runs through an image processing (MUSICA).
The processed images are stored on the internal hard disk of the
NX Workstation. However, the NX Workstation can only provide a limited
amount of storage capacity. The final archiving must be realized either by
printed hardcopies or a PACS.
Moreover, the NX Workstation decodes the routing data and sends the image
data to the selected destinations (e.g. Printer, PACS).
For the NX Workstation a lot of licensed software options (e.g. Interactive
Image Processing, Annotation, Smart Print, etc.) can be purchased.
The CR30-X cannot be run together with ADC QS workstations.
DOCUMENT CONTROL NOTE:
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Chapter 2 / 2
Agfa Company Confidential
CR30-X
Type 5175/100
Edition 1, Revision 0
DD+DIS150.06E
1.4
Functional Description
Shutter
Lock
Image plate
RF-tag
Tray
517511al.cdr
Edition 1, Revision 0
CR30-X
Type 5175/100
Chapter 2 / 3
Agfa Company Confidential
DD+DIS150.06E
Functional Description
Take exposure
5
Take cassette
for next exposure
3
Go to processing station
and select Patient:
The IP* cycle starts.
FireWire cable
517502ah.cdr
Figure 4
Workflow of an examination with the CR30-X system in the view of a
radiologic technologist:
Step 1:
Step 2:
Step 3:
Step 4:
Wait till the IP* cycle is finished. This takes approx. 1 min.
The digitizer sends the image data during scanning (i.e. during
the IP cycle) to the processing station.
Step 5:
Chapter 2 / 4
Agfa Company Confidential
CR30-X
Type 5175/100
Edition 1, Revision 0
DD+DIS150.06E
Functional Description
Chip
Reader
User
Interface
NXApplication
Software
Contr.
E*
E* IP
Handling
Board
E*
Slow Scan
Board
Contr.
PMI Board
(PhotoMultiplier
Interface)
PMT
E*
*E-Label
Slow Scan
Control
TWAIN Interface
Contr.
Optics
PCBs
E*
FireWire
Interface
CR30-X Server
Application
Error
Viewer
CR30-X Service
Client
Calibration
board
Logger
ASAP
Erasure
Board
Digitizer
Control
Image Capture
Control
Tag Reader
Control
Cassette
Handling
Control
IP Handling
Control
Harddisk
517502ag.cdr
NX Workstation
Digitizer
Figure 5
Edition 1, Revision 0
CR30-X
Type 5175/100
Chapter 2 / 5
Agfa Company Confidential
DD+DIS150.06E
Functional Description
Digitizer Boot-Up
3x
1 beep
repeat endless
if required
3
Send Self Test
Signal to all SW modules
Selftest okay?
NO
Error message
in NX Error Viewer.
Boot-up stops
YES
YES
Warning message
in Error Viewer
Remove Cassette
or IP and continue
NO
Motors audible
Cassette or IP
found?
Approx. 50 s.
IP Clean-up
Cycle
YES
Polygon audible
NO
Approx. 10 s.
NX connected?
2
Visible
LED dark
Boot-up of PMI
and all other CAN nodes
Audible
No audible
activities
CR30-X Boot-Up
Digitizer
Switch-On
Approx. 10 s.
NX Viewer
running?
NO
YES
Digitizer idle
LED
green on
YES
No audible
activities
NO
Error message
in NX Error Viewer.
Boot-up stops
Approx. 1 s.
Initialization okay?
517502eb.cdr
Legend:
33 SW Modules:
Figure 6
Chapter 2 / 6
Agfa Company Confidential
CR30-X
Type 5175/100
Edition 1, Revision 0
DD+DIS150.06E
Functional Description
Scanning
Laser Diode
6 Facet Polygon
EOL Sensor
4
5
6
7
8
Geometric Correction
Sensors (1 to 16)
9
10
11
12
13
14
15
16
Calibration Board
BOL Sensor
517502ae.cdr
Figure 7
Edition 1, Revision 0
CR30-X
Type 5175/100
Chapter 2 / 7
Agfa Company Confidential
DD+DIS150.06E
Functional Description
5.1
5.2
Geometric Correction:
Jitter Correction:
To correct irregularities in the polygon run, the digitizer measures the timings
between the BOL and EOL signal during each scan. Certain irregularities,
which may be caused by external vibrations or the polygon ball bearings are
smoothed by calculating mean values of several scan lines.
When an IP is inserted the polygon starts up. The BOL/EOL sensors measure
the position of the polygon mirrors (they even detect each single facet)
When the IP arrives for scanning, the sensors of the calibration board detect
the IP edge and generate the begin of scan signal (BOS).
The begin of line signal (BOL) triggers a timer for switching on the laser beam
and the AD window.
The length of the AD window and "Laser On" depends on the image plate size.
Chapter 2 / 8
Agfa Company Confidential
CR30-X
Type 5175/100
Edition 1, Revision 0
DD+DIS150.06E
Functional Description
BOL
EOL
8 9 10 11 12 13 14 15 16
AD Window
IP Width
517502ac.cdr
Note: Timings are unproportional distorted to show correlation between different signals
Figure 8
5.3
Edition 1, Revision 0
CR30-X
Type 5175/100
Chapter 2 / 9
Agfa Company Confidential
DD+DIS150.06E
Functional Description
Chapter 2 / 10
Agfa Company Confidential
CR30-X
Type 5175/100
Edition 1, Revision 0
DD+DIS150.06E
Functional Description
REF CAL
REF LUM
Shading Calibration
PMT SN
Flash ROM*
EEPROM*
E-Label*
PMT
Connector*
Light Collector
Blue LED
Figure 9
Edition 1, Revision 0
CR30-X
Type 5175/100
Chapter 2 / 11
Agfa Company Confidential
DD+DIS150.06E
Functional Description
Stall Detection
Figure 10
Stall detection =
Motor stall (home/end position) detection without the use of additional
sensors
Made by evaluation of stepper motor control curve on the motor controller
chip
Advantage compared to home/end position detection via light sensors:
Monitoring of motor run not only in home and end position, but also
between home and end position (except acceleration / deceleration time)
Less sensors (costs, reliability) and cabling (costs).
How does it work in general:
When the stepper motor is accelerating the stall detection is switched off,
as torque differs to much during acceleration.
When the motor reaches a stable speed, e.g. after 200 steps, the stall
detection is active for a number of steps, e.g. 5000 steps: The motor
control curve is evaluated by the motor controller chip.
Chapter 2 / 12
Agfa Company Confidential
CR30-X
Type 5175/100
Edition 1, Revision 0
DD+DIS150.06E
Functional Description
When the torque increases and reaches a certain "switch off torque", as
the motor reaches the end position, the motor controller detects the stall
position and drives the motor a certain number of steps (e.g. 50 steps) till
it stops.
During decelration time the stall detection is switched off, as torque differs
to much during deceleration.
Rubber pieces in the end position support that the motor can be driven
with relatively high speed towards the end position.
If the torque increases to the "switch-off" torque before the motor is
supposed to reach the stall position, the motor is also stopped. The
software however "knows" that this stop comes too early and issues an
error message (e.g. "IP jam").
NOTE:
The motor curves shown above are simplified. In reality the motor speed can
be reduced just before the stall position, or can even be reduced in several
steps before it stops.
Edition 1, Revision 0
CR30-X
Type 5175/100
Chapter 2 / 13
Agfa Company Confidential
Chapter 2
HEALTHCARE
Functional Description
Imaging Services
CR 30-X
Software / Hardware Compatibility Chart
Document History
Edition.
Revision
Release
Date
Changes
compared to previous Version 1.2
1.3
04-2009
Referenced Documents
Document
Title
Edition 1, Revision 3
04-2009 printed in Germany
Agfa Company Confidential
1
Functional Description
DD+DIS125.08E
Manufacturer
Agfa HealthCare N.V.
Publisher
Agfa-Gevaert HealthCare GmbH
Tegernseer Landstrae 161
D - 81539 Mnchen
Germany
Copyright 2009 Agfa HealthCare N.V.
All rights reserved.
Technical modifications reserved.
AGFA and the Agfa-Rhombus are trademarks of Agfa HealthCare N.V.
WARNING:
Improper operation or service activities may cause damage or injuries.
INSTRUCTION:
(1)
(2)
Strictly observe all safety directions within the "Generic Safety Directions" and on
the product.
NOTE:
To verify the latest version of single documents and of Service Manuals refer to the
Document Type Order List in the GSO library.
Edition 1, Revision 3
04-2009
CR 30-X
Type 5175 / 100/110
Chapter 2 / 2
Agfa Company Confidential
Functional Description
DD+DIS125.08E
software
hardware
supported applications
Indicates which CR 30-X System components are released for the respective
CR 30-X System Versions.
NOTE:
Edition 1, Revision 3
04-2009
CR 30-X
Type 5175 / 100/110
Chapter 2 / 3
Agfa Company Confidential
Functional Description
DD+DIS125.08E
System Version
002
003
Initial release
Introduction
CRE_1206
Introduction
CRE_1309
yes
no
yes
yes
yes
yes
yes
yes
no
no
no
no
no
no
yes
yes
CRE_1010A
CRE_12061)
CRE_1309
CRE_1309
CRE_1309
CRE_1309A
CRE_1405
NX 1.0.3203 (SP1)
yes
yes
no
no
no
no
no
no
yes
no
no
no
NX 2.0.6805
no
no
no
yes
yes
yes
no
no
NX 2.0.4
no
no
no
no
no
no
yes
yes
NX 2.0.6805
no
no
no
yes
yes
yes
no
no
NX 2.0.6805 (SU1)
no
no
no
no
no
no
yes
yes
NX 2.0.6805 (SU2)
no
no
no
no
no
no
yes
yes
NX 2008 (DELL)
no
no
no
no
no
no
no
NX 2008 (Laptop)
no
no
no
no
no
no
no
no
no
no
no
no
no
no
no
no
no
no
no
no
no
no
no
no
no
no
no
no
no
yes
yes
yes
yes
yes
004
005
005A
009
010
011
012
Introduction
CRE_1509
Introduction
NX2008
Introduction
RAID on
NX2008
Introduction
NX 8000 4)
yes
yes
yes
yes
yes
yes
yes
yes
yes
yes
CRE_1405
CRE_1405
CRE_1509
CRE_1405
CRE_1509
CRE_1405
CRE_1509
CRE_1405
CRE_1509
no
no
no
no
no
no
no
no
no
no
no
no
no
no
no
no
no
no
no
yes
yes
no
no
no
no
no
no
no
no
yes
yes
no
no
no
yes
yes
no
no
no
no
no
no
yes
yes
yes
no
no
no
yes
yes
yes
no
no
no
no
yes
yes
yes
no
no
no
no
yes
yes
yes
no
no
no
no
no
yes
yes
yes
no
no
no
no
no
yes
yes
yes
yes
yes
yes
yes3)
yes3)
yes
yes
yes
3)
3)
006
007
008
Digitizer
FLFS Phase 2 Introduction
5175/110 +
license NX
FLFS phase 2
WHITE
LABEL
DIGITIZER
Digitizer Software
Workstation
NX 8000
4)
yes
yes
yes
yes
yes
yes
yes
yes
yes
yes
yes
yes
yes
NX OPTIVIEW
yes
yes
yes
yes
yes
yes
yes
yes
yes3)
yes3)
yes3)
yes3)
yes3)
NX INTEGRATED WORKFLOW
yes
yes
yes
yes
yes
yes
yes
yes
yes3)
yes3)
yes3)
yes3)
yes3)
3)
3)
3)
3)
NX PRECISION TOOLS
yes
yes
yes
yes
yes
yes
yes
yes
yes
yes
yes
yes
yes3)
NX RIS CONNECTIVITY
yes
yes
yes
yes
yes
yes
yes
yes
yes3)
yes3)
yes3)
yes3)
yes3)
3)
3)
3)
3)
yes3)
3)
yes3)
NX QUALITY ASSURANCE
NX PEDIATRIC
yes
yes
yes
yes
yes
yes
yes
yes
yes
yes
yes
yes
yes
yes
yes
yes
yes
3)
yes
yes
3)
yes
yes
3)
yes
yes
yes
Edition 1, Revision 3
04-2009
CR 30-X
Type 5175 / 100/110
Chapter 2 / 4
Agfa Company Confidential
Functional Description
Software / Hardware Compatibility Chart CR 30-X System
DD+DIS125.08E
System Version
001
002
003
004
005
005A
006
007
008
009
010
011
012
no
no
no
yes
yes
yes
yes
yes
yes3)
yes3)
yes3)
yes3)
yes3)
NX WHITE LABEL
no
no
no
no
no
no
no
no
yes2)
yes2)
yes2)
yes2)
yes2)
NX VETERINARY-EL
no
no
no
no
no
no
no
no
no
no
yes
yes
yes
NX VETERINARY ENTERPRISE
no
no
no
no
no
no
no
no
no
no
no
no
no
no
no
no
no
no
no
no
no
no
no
yes
yes
yes
no
no
no
no
no
no
no
no
no
no
yes
yes
yes
no
no
no
no
no
no
no
no
no
no
no
no
yes
NX Musica2 Neonatal
no
no
no
no
no
no
no
no
no
no
no
no
yes
no
no
no
no
no
no
no
no
no
no
no
no
yes
no
no
no
no
no
no
no
no
no
no
no
no
yes
no
no
no
no
no
no
no
no
no
no
no
no
yes
yes
yes
yes
yes
yes
yes
yes
yes
yes
yes
yes
yes
yes
yes
yes
yes
yes
yes
yes
yes
yes
yes
yes
yes
yes
yes
yes
yes
yes
yes
yes
yes
yes
yes
yes
yes
yes
yes
yes
no
no
no
no
yes
yes
yes
yes
yes
yes
yes
yes
yes
no
no
no
no
yes
yes
yes
yes
yes
yes
yes
yes
yes
Optional CR EasyLift
no
no
no
no
yes
yes
yes
yes
yes
yes
yes
yes
yes
no
no
no
no
no
no
yes
yes
yes
yes
yes
yes
yes
yes
yes
yes
yes
yes
yes
yes
yes
yes
yes
yes
yes
yes
yes
yes
yes
yes
yes
yes
yes
yes
yes
yes
yes
yes
yes
yes
yes
yes
yes
yes
yes
yes
yes
yes
yes
yes
yes
yes
yes
yes
yes
yes
yes
yes
yes
yes
yes
yes
yes
yes
yes
System Options
Cassettes
1)
= CRE_1206 A, C or D
= only available for digitizer type 5175/110.
3)
= only available for digitizer type 5175/100
4)
NX 8000 = NX 2.0.8000/NX 3.0.8000
2)
Edition 1, Revision 3
04-2009
CR 30-X
Type 5175 / 100/110
Chapter 2 / 5
Agfa Company Confidential
DD+DIS125.08E
Functional Description
Software / Hardware Compatibility Chart CR 30-X System
NX 2008 - Service Bulletin no. 02 - RAID is available for NX 2008 on DELL Optiplex
755 Workstations DD+DIS237.08E
(Intranet link / extranet link)
Edition 1, Revision 3
04-2009
CR 30-X
Type 5175 / 100/110
Chapter 2 / 6
Agfa Company Confidential
Chapter 3.1
HEALTHCARE
Imaging Services
CR 30-X
Type 5175 /100/110
Document History
Edition.
Revision
Release
Date
Changes
compared to previous Version 1.0
1.1
04-2009
Referenced Documents
Document
Title
n.a.
n.a.
Edition 1, Revision 1
04-2009 printed in Germany
DD+DIS150.06E
Manufacturer
Agfa HealthCare N.V.
Publisher
Agfa-Gevaert HealthCare GmbH
Tegernseer Landstrae 161
D - 81539 Mnchen
Germany
Copyright 2009 Agfa HealthCare N.V.
All rights reserved.
Technical modifications reserved.
AGFA and the Agfa-Rhombus are trademarks of Agfa HealthCare N.V.
WARNING:
Improper operation or service activities may cause damage or injuries.
INSTRUCTION:
(1)
(2)
Strictly observe all safety directions within the "Generic Safety Directions" and on
the product.
NOTE:
To verify the latest version of single documents and of Service Manuals refer to the
Document Type Order List in the GSO library.
Edition 1, Revision 1
04-2009
CR 30-X
Type 5175 /100/110
Chapter 3.1 / 2
DD+DIS150.06E
LIST OF CONTENTS
3.2
3.3
3.3.1
3.3.2
3.3.3
3.3.4
3.3.5
Edition 1, Revision 1
04-2009
CR 30-X
Type 5175 /100/110
Chapter 3.1 / 3
DD+DIS150.06E
Intended Use
This device must only be used to scan exposed X-ray cassettes, containing an
erasable image plate (IP). The digitizer is part of a system, consisting of X-ray
cassettes with erasable phosphor image plates and a workstation where the X-ray
cassettes are identified and the resulting digital image information is further processed
and routed. It is intended that this device is only operated in a radiological environment
by qualified staff.
NOTE:
One sticker on the front cover indicates the
correct cassette positioning when inserting the
cassette.
517631ap.cdr
Figure 2
Edition 1, Revision 1
04-2009
CR 30-X
Type 5175 /100/110
Chapter 3.1 / 4
DD+DIS150.06E
This device complies with Part 15 of the FCC rules. Operation is subject to the following two conditions:
(1) This device may not cuase harmful interference (2) This device must accept any interference
received, including interference that may cause undesired operation.
This Class A digital apparatus meets all requirements of the Canadian Interference Causing Equipment
Regulations.
Cet appareil numerique de la classe A respecte toutes les exigences du Reglement sur le material
brouillant du Canada.
Type B equipment
Gerte Typ B
Appareil du type B
Thi s device complies with Part 15 of the FCC rules. Operation is subject to the following two conditions:
(1) This device may not cuase harmful interference (2) This device must accept any interference
received, including interference that may cause undesired operation.
Thi s Class A digital apparatus meets all requirements of the Canadian Interference Causi ng Equipment
Regulations.
Cet appareil numerique de la classe A respecte toutes les exigences du Reglement sur le material
brouillant du Canada.
Type B equipment
Gerte Typ B
Appareil du type B
M anufacturer
ME DI CA LE LE CTR C
I AL
EQ UI PM EN T
WITH R ESP EC T TO
ELE CTR IC S HO C K, FI RE ,
AN D ME CHA N C
I AL
HA ZAR DS O N LYI N
AC CO RDA NC E WI TH
U L60601- 1 /
CA N/ CSA C 22. 2N O .601 .1
9F 03
U
R
US
Mad einGermany
Peissenberg
MMMMM-YYYY
FireWire
Manufacturer
517531ae.cdr
MEDICAL ELECTRICAL
EQUIPMENT
WITH RESPECT TO
ELECTRIC SHOCK, FIRE,
AND MECHANICAL
HAZARDS ONLY IN
ACCORDANCE WITH
UL60601-1 /
CAN/CSA C22.2 NO.601.1
9F03
Connector
C
U
R
US
Made in Germany
Peissenberg
MMMMM-YYYY
Direction
FireWire
NOTE:
The type label is located in the rear. It shows
Manufacturer
Type number
Serial number
Input voltage, current and frequency
Production date
Production site
Information about certificates (CE label
etc.)
WARNING:
Risk of electrical shock if supplementary
earth connector is disconnected with one
hand and the other hand touches a ground
connection (e.g. rear cover of the digitizer).
Manufacturer
MEDICAL ELECTRICAL
EQUIPMENT
WITH RESPECT TO
ELECTRIC SHOCK, FIRE,
AND MECHANICAL
HAZARDS ONLY IN
ACCORDANCE WITH
UL60601-1 /
CAN/CSA C22.2 NO.601.1
9F03
U
R
US
Made in Germany
Peissenberg
MMMMM-YYYY
Figure 4
Figure 5
Edition 1, Revision 1
04-2009
CR 30-X
Type 5175 /100/110
Chapter 3.1 / 5
DD+DIS150.06E
NOTE:
With covers closed the digitizer is a Class 1
laser product and not dangerous to the human
eye or skin.
Figure 6
CAUTION:
Risk of damage of the PMI board if FireWire
connector is inserted wrong.
Insert the FireWire connector in the rear as
indicated on the sticker.
Connector
Direction
FireWire
Figure 7
WARNING:
The erasure unit is hot - up to 200 C: Risk
of burns.
Avoid contact with the Erasure Unit.
Observe the relevant sticker on the Erasure
Unit.
Figure 8
WARNING:
Improper ground connections inside the
device may lead to electric shocks.
Make sure, that the grounding connection to
the digitizer frame is present.
Figure 9
Edition 1, Revision 1
04-2009
CR 30-X
Type 5175 /100/110
Chapter 3.1 / 6
DD+DIS150.06E
WARNING:
80 mW Laser (Class 3B) in the OPTIC MODULE may cause eye injuries.
Avoid direct and indirect eye contact.
Do not keep tools in the laser beam when the device is switched on or the
CASSETTE UNIT OPEN SAFETY SWITCH is overridden with a service key: Risk of
reflection.
Do not open the cover of the OPTIC MODULE.
OPTICS CENTER
DO NOT OPEN
No service / maintenance
parts inside
OPTICS CENTER
DO NOT OPEN
No service / maintenance
parts inside
DANGER
LASER LIGHT
AVOID DIRECT EXPOSURE
TO BEAM
DIODE LASER
120 mW MAX OUTPUT
at 640-670 nm
CLASS IIIb LASER PRODUCT
Figure 10
Edition 1, Revision 1
04-2009
CR 30-X
Type 5175 /100/110
Chapter 3.1 / 7
DD+DIS150.06E
3.1
WARNING:
Excessive vibrations during scanning may decrease image quality.
The structure and stability of the table used, need to be suitable in relation with the size
and weight of the system. The table may not be subject to excessive shock and
vibrations from other sources.
WARNING:
The digitizer uses high voltage. Electrical shock possible.
A ground fault interrupter (GFI) (IN = 30 mA) must be installed at a suitable point in
the circuit (in compliance with VDE 664).
Ensure that there is either a mains plug or an all-cable disconnecting device in the
internal installation fitted near the digitizer and that it is easily accessible.
WARNING:
Vibrations caused by transport in a truck (mobile usage) or earthquakes may
cause that the digitizer falls off the table.
During transport in a truck for mobile usage and at installations in earthquake areas the
digitizer has to be secured by the optional Wall Fixation Kit. See chapter 11, Installation
Planning, section "Accessories".
Edition 1, Revision 1
04-2009
CR 30-X
Type 5175 /100/110
Chapter 3.1 / 8
DD+DIS150.06E
3.2
3.3
3.3.1
CAUTION:
Sharp edges inside the digitizer: Cut or abrasion possible.
Be careful at maintenance and replacement of parts.
CAUTION:
Risk of injuries if the cassette unit open
safety switch is overridden with a service key.
C a s s e t t e U n it o p e n
S a f e t y S w it c h ( S 2 )
5 1 7 5 3 1 a b .c d r
Figure 11
Edition 1, Revision 1
04-2009
CR 30-X
Type 5175 /100/110
Chapter 3.1 / 9
DD+DIS150.06E
3.3.2
3.3.3
3.3.4
Edition 1, Revision 1
04-2009
CR 30-X
Type 5175 /100/110
Chapter 3.1 / 10
DD+DIS150.06E
3.3.5
3.3.5.1
Task
Section Reference
3.3.5.1
3.3.5.2
200
517511ai.cdr
Figure 12
Edition 1, Revision 1
04-2009
CR 30-X
Type 5175 /100/110
Chapter 3.1 / 11
DD+DIS150.06E
517509ah.cdr
Figure 13
IMPORTANT:
In case the electrical resistance of one measurement exceeds the limits (0.2 Ohm
without mains cable; 0.3 Ohm with mains cable) check the digitizer for loose grounding
connections!
3.3.5.2
Edition 1, Revision 1
04-2009
CR 30-X
Type 5175 /100/110
Chapter 3.1 / 12
Chapter 3.2
HEALTHCARE
Imaging Services
CR 30-X
Type 5175 / 100/110
Document History
Edition.
Revision
Release
Date
Changes
compared to previous Version 1.0:
1.1
04-2008
Referenced Documents
Document
Title
n.a.
n.a.
Edition 1, Revision 1
04-2008 printed in Germany
DD+DIS150.06E
Manufacturer
Agfa HealthCare N.V.
Publisher
Agfa-Gevaert HealthCare GmbH
Tegernseer Landstrae 161
D - 81539 Mnchen
Germany
Copyright 2008 Agfa HealthCare N.V.
All rights reserved.
Technical modifications reserved.
AGFA and the Agfa-Rhombus are trademarks of Agfa HealthCare N.V.
WARNING:
Improper operation or service activities may cause damage or injuries.
INSTRUCTION:
(1)
(2)
Strictly observe all safety directions within the "Generic Safety Directions" and on
the product.
NOTE:
To verify the latest version of single documents and of Service Manuals refer to the
Document Type Order List in the GSO library.
Edition 1, Revision 1
04-2008
CR 30-X
Type 5175 / 100/110
Chapter 3.2 / 2
DD+DIS150.06E
LIST OF CONTENTS
Required Tools..........................................................................................................................4
1.2
Recommended Tools................................................................................................................5
2.1.1
2.2
2.2.1
2.2.2
2.2.3
2.2.4
2.2.5
2.2.6
2.2.7
2.2.8
2.2.9
2.2.10
2.2.11
2.2.12
2.2.13
2.2.14
2.2.15
2.2.16
2.2.17
2.2.18
Edition 1, Revision 1
04-2008
CR 30-X
Type 5175 / 100/110
Chapter 3.2 / 3
DD+DIS150.06E
1.1
Required Tools
In addition to the conventional service tools which every service engineer has in his
toolbox, the following tools are required for service on the digitizer:
Tool
Part Number
Purpose
CR Screen Cleaner
(4 bottles)
Service key
Vacuum Cleaner
Commercially available.
See also note next page.
NOTE:
Agfa has a vacuum cleaner in its service spare parts assortment:
Description
Order number
Edition 1, Revision 1
04-2008
CR 30-X
Type 5175 / 100/110
Chapter 3.2 / 4
DD+DIS150.06E
1.2
Recommended Tools
Tool
Part Number
Purpose
Canned air
Service PC
Commercially
available.
NOTE:
As some screws are difficult accessible, it is
recommended to have a 3 mm hex key with
ball-shaped head and a 6 mm magnetic hex
key in the tools assortment.
3 mm,
with ball-shaped head
6 mm (magnetic)
Figure 1
Edition 1, Revision 1
04-2008
CR 30-X
Type 5175 / 100/110
Chapter 3.2 / 5
DD+DIS150.06E
Edition 1, Revision 1
04-2008
CR 30-X
Type 5175 / 100/110
Chapter 3.2 / 6
DD+DIS150.06E
NOTE:
It is possible to use the CR30-X service client also on the service PC.
Following functions can be accessed:
Directly after installation:
Function Explain Error Code
After Restore of a CR30-X backup file:
View Infocounter
View E-Label entries
View Service Activity Log
View Error Statistics
Prerequisites for Installation:
Windows XP (home or professional)
.Net Visual J# (to be obtained via www.microsoft.com)
2.1
All steps for software upgrade / update of the digitizer are described in a
document which is part of the software on MEDNET (readme or enclosure
document). Depending on the software version, the procedure might be
different to the steps described here.
IMPORTANT:
Before starting any software de-installation activity at the control PC, switch off the
digitizer. Otherwise the digitizer driver on the control PC will automatically
be de-installed.
(1)
Select "START - Log Off <current user>" to log-off the current user.
(2)
(3)
Edition 1, Revision 1
04-2008
CR 30-X
Type 5175 / 100/110
Chapter 3.2 / 7
DD+DIS150.06E
(4)
Figure 3
(5)
(6)
Deinstall the digitizer software on the control PC: Select "Start Settings
Control Panel Add or Remove Programs CR 30-X".
(7)
If at the end of the deinstallation process of the old software a window pops up,
telling: "You must restart your system ...." continue with section 2.1.1 next page.
Otherwise continue with next step.
(8)
Copy file "Setup CR 30-X_xxxx.msi" (xxxx = software version) from USB stick or
CD ROM to drive D: on the control PC.
(9)
Figure 4
(11) When finished, restart the control PC.
DOCUMENT CONTROL NOTE:
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 1
04-2008
CR 30-X
Type 5175 / 100/110
Chapter 3.2 / 8
DD+DIS150.06E
2.1.1
Figure 5
Defect
The NX software or the error viewer have not been closed before
software de-installation.
Solution
Symptom 2
After reboot of the control PC the screen "Welcome to the Found New
Hardware Wizard" appears.
Figure 6
Edition 1, Revision 1
04-2008
CR 30-X
Type 5175 / 100/110
Chapter 3.2 / 9
DD+DIS150.06E
Defect
Solution
(1)
(2)
Install the driver via "found new hardware wizard" which pops up
automatically after software upload to the digitizer (and restart of
the control PC). See Figure 7.
Figure 7
Edition 1, Revision 1
04-2008
CR 30-X
Type 5175 / 100/110
Chapter 3.2 / 10
DD+DIS150.06E
2.2
Menu
Refer to ...
Section 2.2.1
E-Label
Section 2.2.2
SW-Versioning
Section 2.2.3
Reporting
Service Activity Log
Section 2.2.4
Error Statistics
Section 2.2.5
HOT Report
Section 2.2.6
Modification
SW Upload
Section 2.2.7
Section 2.2.8
Section 2.2.9
Section 2.2.10
Selftest of Device
Section 2.2.11
Preventive Maintenance
Calibration (Shading Calibration)
Section 2.2.12
Stall Calibration
Section 2.2.13
Configuration
Site specific data
Section 2.2.14
Section 2.2.15
Exchanged components
Section 2.2.16
Maintenance Indicator
Section 2.2.17
Section 2.2.18
Edition 1, Revision 1
04-2008
CR 30-X
Type 5175 / 100/110
Chapter 3.2 / 11
DD+DIS150.06E
2.2.1
PURPOSE
It facilitates recognition of failures and allows reducing the repair time in the field or the
repair center.
USAGE
Figure 8
Edition 1, Revision 1
04-2008
CR 30-X
Type 5175 / 100/110
Chapter 3.2 / 12
DD+DIS150.06E
2.2.2
PURPOSE
USAGE
Figure 9
Edition 1, Revision 1
04-2008
CR 30-X
Type 5175 / 100/110
Chapter 3.2 / 13
DD+DIS150.06E
2.2.3
PURPOSE
USAGE
Node
Installed Version
CRE_1509/1096
CRE_1509
CRE_1509
PMI Board/FPGA
HDL Board
SSC Board
Version on PC
Status
ok
CRE_1509/1096
ok
CRE_1509
ok
CRE_1509
NX Workstation
(Control PC)
Digitizer
5175_chapter32_001.cdr
Figure 10
IMPORTANT:
The installed versions on the PMI board, IP Handling board (HDL Board) and Slow
Scan Board (SSC Board) have always to be the same.
Edition 1, Revision 1
04-2008
CR 30-X
Type 5175 / 100/110
Chapter 3.2 / 14
DD+DIS150.06E
2.2.4
PURPOSE
USAGE
Figure 11
NOTE:
When the logfile exceeds 1 MB, a, new, empty ServiceActivity.log file is created.
The previous one is automatically renamed to ServiceActivity1.log (stored in directory
C:\Documents and Settings\All Users\Application Data\Agfa\CR 30-X\Log).
All ServiceActivity.log files are part of a backup.
Edition 1, Revision 1
04-2008
CR 30-X
Type 5175 / 100/110
Chapter 3.2 / 15
DD+DIS150.06E
2.2.5
PURPOSE
USAGE
The error statistics menu shows the occurrence frequency for all errors.
It is used for troubleshooting.
(1) Select "Reporting - Error Statistics"
Figure 12
Edition 1, Revision 1
04-2008
CR 30-X
Type 5175 / 100/110
Chapter 3.2 / 16
DD+DIS150.06E
2.2.6
PURPOSE
The menu Hot Report gives a summary of the last cassette cycle. The most important
events that happened in the digitizer during this cycle are reported.
It is used as a proof during installation, that the digitizer performed a successful IP cycle.
The HOT report is overwritten with each new scan cycle.
USAGE
Figure 13
Edition 1, Revision 1
04-2008
CR 30-X
Type 5175 / 100/110
Chapter 3.2 / 17
DD+DIS150.06E
2.2.7
PURPOSE
After replacement of the PMI board, slow scan board or IP handling board.
NOTE:
All steps for software upgrade / update of the digitizer are described in a document which
is part of the software on MEDNET (readme or enclosure document). Depending on the
software version, the procedure might be different to the steps described here.
USAGE
(1)
(2)
Node
PMI Board
HDL Board
SSC Board
Status
ok
ok
ok
Select All
Clear Selection
NX Workstation
(Control PC)
Digitizer
517501da.cdr
Figure 14
Edition 1, Revision 1
04-2008
CR 30-X
Type 5175 / 100/110
Chapter 3.2 / 18
DD+DIS150.06E
(3)
(4)
(5)
Wait for approx. 7 minutes till message "versions are current" appears in the
message box.
(6)
(7)
Answer the dialogue popping up did you perform any repair or maintenance
activities with "Yes".
(8)
(9)
IMPORTANT:
The installed versions on the PMI board, IP Handling board (HDL Board) and Slow
Scan Board (SSC Board) have always to be the same.
Edition 1, Revision 1
04-2008
CR 30-X
Type 5175 / 100/110
Chapter 3.2 / 19
DD+DIS150.06E
2.2.8
PURPOSE
USAGE
Figure 15
DOCUMENT CONTROL NOTE:
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 1
04-2008
CR 30-X
Type 5175 / 100/110
Chapter 3.2 / 20
DD+DIS150.06E
2.2.9
PURPOSE
USAGE
Figure 16
Edition 1, Revision 1
04-2008
CR 30-X
Type 5175 / 100/110
Chapter 3.2 / 21
DD+DIS150.06E
2.2.10
PURPOSE
USAGE
Figure 17
Edition 1, Revision 1
04-2008
CR 30-X
Type 5175 / 100/110
Chapter 3.2 / 22
DD+DIS150.06E
2.2.11
PURPOSE
USAGE
The Selftest of Device menu allows to reset the digitizer via CR30-X service client. The
functional sequence of the selftest is identical to the selftest during switch-on.
Refer to chapter 2, Functional Description.
(1) Select "Analysis & Repair - Selftest of device"
(2) Wait till the digitizer has rebooted - takes approx. 1 min.
Figure 18
Edition 1, Revision 1
04-2008
CR 30-X
Type 5175 / 100/110
Chapter 3.2 / 23
DD+DIS150.06E
2.2.12
PURPOSE
USAGE
Figure 19
Edition 1, Revision 1
04-2008
CR 30-X
Type 5175 / 100/110
Chapter 3.2 / 24
DD+DIS150.06E
2.2.13
PURPOSE
USAGE
Figure 20
NOTE:
For troubleshooting of a module the final message parameter iRunCal can be used
(see Figure 20):
If executed three consecutive times, the displayed iRunCal value may vary 2 digits
only (e.g. iRunCal values 56, 58, 60). Larger deviations show mechanical problems in the
mechanics of the respective drive.
Edition 1, Revision 1
04-2008
CR 30-X
Type 5175 / 100/110
Chapter 3.2 / 25
DD+DIS150.06E
2.2.14
PURPOSE
USAGE
Menu site specific data allows entering the site specific data for exact identification of
the digitizer. The data are part of the infocounter and the E-labels.
(1) Select "Configuration - Site specific data"
(2) Enter the required data
(3) Select "Update Site Specific Data"
Figure 21
Edition 1, Revision 1
04-2008
CR 30-X
Type 5175 / 100/110
Chapter 3.2 / 26
DD+DIS150.06E
2.2.15
PURPOSE
USAGE
Menu Device specific data allows to read-out the device specific data which were
configured in production.
(1) Select "Configuration - Device specific data".
(2) Select "Read Data".
Figure 22
NOTE:
The Device specific Data are read-only.
Edition 1, Revision 1
04-2008
CR 30-X
Type 5175 / 100/110
Chapter 3.2 / 27
DD+DIS150.06E
2.2.16
PURPOSE
USAGE
Type of Exchange
Module
Board
Module
IP handling board
Board
Module
Optic module
Module
PMI2 board
CM+9.5175.2180.X
Board
Figure 23
Edition 1, Revision 1
04-2008
CR 30-X
Type 5175 / 100/110
Chapter 3.2 / 28
DD+DIS150.06E
NOTE:
Background information to function Configuration Exchanged Components:
For exchange of spare parts with E-Labels 1 we distinguish two scenarios:
Scenario 1
Type of spare part with ELabel
IP handling board
Scenario 2
Spare part is an adjusted
module:
Optic Module
9999
Board
Module
Required selection in
menu Configuration
Exchanged Components
E-Labels contain spare part specific data, entered in production. Following boards have E-Labels included:
PMT module, Printed circuit board in Optic Module, IP handling board, Slow scan board, PMI / PMI2 board.
DOCUMENT CONTROL NOTE:
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 1
04-2008
CR 30-X
Type 5175 / 100/110
Chapter 3.2 / 29
DD+DIS150.06E
2.2.17
PURPOSE
USAGE
Figure 24
Edition 1, Revision 1
04-2008
CR 30-X
Type 5175 / 100/110
Chapter 3.2 / 30
DD+DIS150.06E
2.2.18
PURPOSE
IMPORTANT:
The device specific data of the digitizer are stored in a database in a certain directory
on the control PC and in E-labels of the digitizer.
If these data are lost (e.g. after exchange of the Control PC; Clean Install Procedure
etc.) all data in the digitizer E-labels are copied to the database of the control PC.
Some data like activity log; site specific data and several infocounter data are not
stored on the digitizer and are lost without an up-to-date backup. These data are
however not relevant for proper operation.
Without restore the digitizer will boot-up with an error message 4637 however,
indicating that the serial number of the digitizer is not known.
To have all data as up-to-date as possible after exchange of the Control PC or Clean
Install Procedure follow the basic rules:
Always make a backup after each service intervention unless only data have
been retrieved (e.g. viewing infocounter)
Edition 1, Revision 1
04-2008
CR 30-X
Type 5175 / 100/110
Chapter 3.2 / 31
DD+DIS150.06E
Backup:
(1) Select "Backup & Restore".
(2) Select "Backup selected files" to perform a backup.
Restore:
(1) Select "Backup & Restore".
(2) Select "Restore selected files" to perform a restore.
Figure 25
Edition 1, Revision 1
04-2008
CR 30-X
Type 5175 / 100/110
Chapter 3.2 / 32
DD+DIS150.06E
Remote Desktop
PURPOSE
USAGE
Remote connection to the processing station and so to the digitizer is performed via
Remote Desktop connection.
(1) Connect the service PC to the network where the processing station with digitizer is
located (e.g. via Agfa SRSS).
(2) On the local PC select Start Programs Accessories Communications
Remote Desktop.
(3) Enter the hostname or IP address of the processing station.
(4) Select Connect.
(5) On the processing station start the CR30-X service client.
Figure 26
NOTE:
The processing station is by default enabled for remote desktop. Enabling for
remote desktop is done via function START Control Panel System Remote.
If the CR30-X service client is started in a remote session, following functions are
not accessible, as they require presence of a service engineer at the digitizer:
o Diagnose Cycle
o Endurance run cycle
o Calibration (= Shading Calibration)
Edition 1, Revision 1
04-2008
CR 30-X
Type 5175 / 100/110
Chapter 3.2 / 33
Chapter 3.3
HEALTHCARE
Imaging Services
Troubleshooting
CR 30-X
Type 5175 / 100/110
Please note
Edition 1, Revision 1
07-2008 printed in Germany
Agfa Company Confidential
DD+DIS150.06E
Edition 1, Revision 1
07-2008
CR 30-X
Type 5175 / 100/110
Chapter 3.3 / 2
Agfa Company Confidential
Chapter 3.4
HEALTHCARE
Imaging Services
CR 30-X
Type 5175 / 100/110
Document History
Edition.
Revision
Release
Date
Changes
compared to previous Version 1.0:
1.1
04-2008
Referenced Documents
Document
Title
DD+DIS063.08E
Edition 1, Revision 1
04-2008 printed in Germany
DD+DIS150.06E
Manufacturer
Agfa HealthCare N.V.
Publisher
Agfa-Gevaert HealthCare GmbH
Tegernseer Landstrae 161
D - 81539 Mnchen
Germany
Copyright 2008 Agfa HealthCare N.V.
All rights reserved.
Technical modifications reserved.
AGFA and the Agfa-Rhombus are trademarks of Agfa HealthCare N.V.
WARNING:
Improper operation or service activities may cause damage or injuries.
INSTRUCTION:
(1)
(2)
Strictly observe all safety directions within the "Generic Safety Directions" and on
the product.
NOTE:
To verify the latest version of single documents and of Service Manuals refer to the
Document Type Order List in the GSO library.
Edition 1, Revision 1
04-2008
CR 30-X
Type 5175 / 100/110
Chapter 3.4 / 2
DD+DIS150.06E
LIST OF CONTENTS
Edition 1, Revision 1
04-2008
CR 30-X
Type 5175 / 100/110
Chapter 3.4 / 3
DD+DIS150.06E
+3,3V
+3,3V
+ 1,5V LD2/WATCHDOG
+24V_POL
FPGA_OK
+24V_POL
FPGA_OK
FWI_orange
+ 1,5V
+24V_POL
+12V_OPT
-12V_OPT
+20V_PMT
-10V_PMT
+8V_ANALOG
+12V_CAN
+8V_DISP
S4
+5V_PMT
ON
ON
S3
S1
+ 5V
+ 5V
LD2/WATCHDOG
-12V_OPT
+12V_CAL
ON
S2
/CS1
/CS0
/CS2
/FPGA_TEST4
/CS3
/M23_RESERVE1
/FPGA_TEST3
/M32_RESERVE2
/RESET
S4
S3
ON
ON
1 2
S1
5175_chapter3.4_001.cdr
** +24V_POL
ON
/CS0
/CS1
/CS2
/CS3
/FPGA_TEST4
/FPGA_TEST3
/M23_RESERVE1
/M32_RESERVE2
/RESET
S2
+12V_OPT
-12V_OPT
+12V_CAL
-12V_OPT
+20V_PMT
-10V_PMT
+8V_ANALOG
+12V_CAN
+8V_DISP
** +5V_PMT
Figure 1
LED
Meaning
FPGA_OK
+1,5 V
LD2 Watchdog
+5V
Edition 1, Revision 1
04-2008
CR 30-X
Type 5175 / 100/110
Chapter 3.4 / 4
DD+DIS150.06E
LED
Meaning
/Reset
/M32_Reserve2
/M32_Reserve1
/FPGA_Test3
/FPGA_Test4
/CS3
/CS2
/CS1
/CS0
+5 V PMT
+8 V DISP
+12 V CAN
+8 V Analog
-10 V PMT
+20 V PMT
-12 V CAL
+12 V CAL
-12 V OPT
+12 V OPT
+24 V POL
+3,3 V
FWI_Orange
FWI_Green
*If these LEDs are off, this does not necessarily mean malfunction of the board. May depend on SW version,
how LED is programmed.
DOCUMENT CONTROL NOTE:
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 1
04-2008
CR 30-X
Type 5175 / 100/110
Chapter 3.4 / 5
DD+DIS150.06E
Switch
Meaning
S1 / 1
S1 / 2
S2 / 1
S2 / 2
S3 / 1
S3 / 2
Edition 1, Revision 1
04-2008
CR 30-X
Type 5175 / 100/110
Chapter 3.4 / 6
DD+DIS150.06E
ON
S2
LD2
T1A
F1
T1A
F1#1
Figure 2
LED
Meaning
Red LD2
Switch
Meaning
S1 / 1
S1 / 2
Not used.
S2
IMPORTANT:
Fuses are displayed only to check fuses in case of a suspected defective fuse.
Always replace the board if a fuse is defective.
DOCUMENT CONTROL NOTE:
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 1
04-2008
CR 30-X
Type 5175 / 100/110
Chapter 3.4 / 7
DD+DIS150.06E
T3.5A
F7
F10
T1A
T1A
F1#2
F1#1
T500mA
T1A
F8
LD1
ON
5175_chapter3.4_002.cdr
S1
ON
1 2
S2
Figure 3
LED
Meaning
Red LD1
Switch
Meaning
S1
S2 / 1
S2 / 2
Not used.
IMPORTANT:
Fuses are displayed only to check fuses in case of a suspected defective fuse.
Always replace the board if a fuse is defective.
DOCUMENT CONTROL NOTE:
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 1
04-2008
CR 30-X
Type 5175 / 100/110
Chapter 3.4 / 8
DD+DIS150.06E
F15
T3.5A
F16
T2.5A
T2.5A
F1#1
F20
T2A
F27
5175_chapter3.4_004.cdr
T5A
F13
F7A
F25
T1A
F23
T1A
F17
T1A
F24
T1A
F19
T1A
F18
T2.5A
F1
Figure 4
IMPORTANT:
Fuses are displayed only to check fuses in case of a suspected defective fuse.
Always replace the board if a fuse is defective.
Edition 1, Revision 1
04-2008
CR 30-X
Type 5175 / 100/110
Chapter 3.4 / 9
DD+DIS150.06E
blue
green or
red
5175_chapter3.4_005.cdr
Figure 5
LED
Meaning
Blue
Green
ON if
Ready / Stand-by
Green flashing
Red
Red flashing
ON during .
Boot-up
Error
Edition 1, Revision 1
04-2008
CR 30-X
Type 5175 / 100/110
Chapter 3.4 / 10
Chapter 3.5
HEALTHCARE
Imaging Services
CR 30-X
Type 5175 / 100/110
Document History
Edition.
Revision
Release
Date
Changes
compared to previous Version 1.1
1.2
03-2009
Referenced Documents
Document
Title
DD+DIS150.06E
Edition 1, Revision 2
03-2009 printed in Germany
DD+DIS150.06E
Manufacturer
Agfa HealthCare N.V.
Publisher
Agfa-Gevaert HealthCare GmbH
Tegernseer Landstrae 161
D - 81539 Mnchen
Germany
Copyright 2009 Agfa HealthCare N.V.
All rights reserved.
Technical modifications reserved.
AGFA and the Agfa-Rhombus are trademarks of Agfa HealthCare N.V.
WARNING:
Improper operation or service activities may cause damage or injuries.
INSTRUCTION:
(1)
(2)
Strictly observe all safety directions within the "Generic Safety Directions" and
on the product.
NOTE:
To verify the latest version of single documents and of Service Manuals refer to the
Document Type Order List in the GSO library.
Edition 1, Revision 2
03-2009
CR 30-X
Type 5175 / 100/110
Chapter 3.5 / 2
DD+DIS150.06E
LIST OF CONTENTS
1.2
1.3
1.4
2.2
2.3
2.4
2.5
2.6
2.7
2.8
2.9
2.10
2.11
2.12
Replacing the Side Plate Transport Unit (left Hand Side) ......................................................43
2.13
2.14
2.15
2.16
2.17
2.18
2.19
2.20
2.21
2.22
2.23
Edition 1, Revision 2
03-2009
CR 30-X
Type 5175 / 100/110
Chapter 3.5 / 3
DD+DIS150.06E
2.24
2.25
3.2
3.3
3.4
3.5
3.6
3.7
3.8
3.9
4.2
4.3
Edition 1, Revision 2
03-2009
CR 30-X
Type 5175 / 100/110
Chapter 3.5 / 4
DD+DIS150.06E
1.1
Figure 1
REQUIRED TOOLS:
3 mm
REQUIRED TIME:
10 min.
DISASSEMBLY:
(1)
(2)
(3)
(4)
B
A
A
Figure 2
Edition 1, Revision 2
03-2009
CR 30-X
Type 5175 / 100/110
Chapter 3.5 / 5
DD+DIS150.06E
(5)
Figure 3
(6)
5175R3-5_004.cdr
Figure 4
(7)
(8)
Figure 5
(9)
Figure 6
DOCUMENT CONTROL NOTE:
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 2
03-2009
CR 30-X
Type 5175 / 100/110
Chapter 3.5 / 6
DD+DIS150.06E
Figure 7
ASSEMBLY:
(1) Mount in reverse order.
Verification:
Result:
1.2
Figure 8
REQUIRED TOOLS:
3 mm,
with ball-shaped head
DOCUMENT CONTROL NOTE:
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 2
03-2009
CR 30-X
Type 5175 / 100/110
Chapter 3.5 / 7
DD+DIS150.06E
REQUIRED TIME:
10 min.
DISASSEMBLY:
(1)
(2)
(3)
(4)
B
A
A
Figure 9
(3)
(4)
Figure 10
(5)
(6)
Figure 11
Edition 1, Revision 2
03-2009
CR 30-X
Type 5175 / 100/110
Chapter 3.5 / 8
DD+DIS150.06E
ASSEMBLY:
(1)
Verification:
Result:
1.3
Figure 12
REQUIRED TOOLS:
3 mm,
with ball-shaped head
REQUIRED TIME:
10 min.
Edition 1, Revision 2
03-2009
CR 30-X
Type 5175 / 100/110
Chapter 3.5 / 9
DD+DIS150.06E
DISASSEMBLY:
(1)
(2)
(3)
(4)
B
A
A
Figure 13
(5)
(6)
Figure 14
(7)
(8)
Figure 15
ASSEMBLY:
(1)
Edition 1, Revision 2
03-2009
CR 30-X
Type 5175 / 100/110
Chapter 3.5 / 10
DD+DIS150.06E
Verification:
Result:
1.4
(1)
(2)
Figure 16
REQUIRED TOOLS:
3 mm,
with ball-shaped head
REQUIRED TIME:
5 min.
DISASSEMBLY:
(1)
(2)
(3)
(4)
B
A
A
Figure 17
DOCUMENT CONTROL NOTE:
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 2
03-2009
CR 30-X
Type 5175 / 100/110
Chapter 3.5 / 11
DD+DIS150.06E
(5)
(6)
Figure 18
ASSEMBLY:
(1) Mount in reverse order.
Verification:
Result:
Edition 1, Revision 2
03-2009
CR 30-X
Type 5175 / 100/110
Chapter 3.5 / 12
DD+DIS150.06E
2.1
Figure 19
REQUIRED TOOLS:
3 mm,
with ball-shaped head
2 mm
Medium size
REQUIRED TIME:
90 min.
DISASSEMBLY:
(1)
(2)
(3)
(4)
B
A
A
Figure 20
Edition 1, Revision 2
03-2009
CR 30-X
Type 5175 / 100/110
Chapter 3.5 / 13
DD+DIS150.06E
(5)
Figure 21
(6)
5175R3-5_004.cdr
Figure 22
(7)
(8)
Figure 23
(9)
Figure 24
DOCUMENT CONTROL NOTE:
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 2
03-2009
CR 30-X
Type 5175 / 100/110
Chapter 3.5 / 14
DD+DIS150.06E
Figure 25
(12) Take out cleaning brush holder.
Figure 26
(13) Remove screw at erasure unit.
(14) Disconnect two cables at erasure unit.
(15) Take out erasure unit.
Figure 27
CAUTION:
IP jams possible if IP guide plate is deformed.
Do not bend IP guide plate at removal.
Edition 1, Revision 2
03-2009
CR 30-X
Type 5175 / 100/110
Chapter 3.5 / 15
DD+DIS150.06E
Figure 28
(17) Disconnect cable at roller IP sensor.
Figure 29
(18) Open three cable holders at right hand
side.
Figure 30
(19) Open cable holders at left hand side.
Figure 31
DOCUMENT CONTROL NOTE:
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 2
03-2009
CR 30-X
Type 5175 / 100/110
Chapter 3.5 / 16
DD+DIS150.06E
Figure 32
(21) Cut one cable tie at upper part transport
unit cable harness left hand side.
Figure 33
(22) Unplug P4, P17, P25 and P29 at IP
handling board.
Figure 34
(23) Remove 2 screws of upper part transport
unit right hand side.
Figure 35
Edition 1, Revision 2
03-2009
CR 30-X
Type 5175 / 100/110
Chapter 3.5 / 17
DD+DIS150.06E
Figure 36
(25) Take off upper part transport unit.
Figure 37
(26) With a screwdriver remove two ball
bearings at transport unit left hand side.
Figure 38
Figure 39
Edition 1, Revision 2
03-2009
CR 30-X
Type 5175 / 100/110
Chapter 3.5 / 18
DD+DIS150.06E
Figure 40
Figure 41
ASSEMBLY:
(1)
Verification:
Result:
Edition 1, Revision 2
03-2009
CR 30-X
Type 5175 / 100/110
Chapter 3.5 / 19
DD+DIS150.06E
2.2
Figure 42
REQUIRED TOOLS:
3 mm
1,5 mm
REQUIRED TIME:
30 min.
DISASSEMBLY:
(1)
(2)
(3)
(4)
B
A
A
Figure 43
(5)
Figure 44
DOCUMENT CONTROL NOTE:
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 2
03-2009
CR 30-X
Type 5175 / 100/110
Chapter 3.5 / 20
DD+DIS150.06E
(6)
5175R3-5_004.cdr
Figure 45
(7)
(8)
Figure 46
(9)
Figure 47
(10) Disconnect ground cable at right hand side.
(11) Remove front cover.
Figure 48
DOCUMENT CONTROL NOTE:
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 2
03-2009
CR 30-X
Type 5175 / 100/110
Chapter 3.5 / 21
DD+DIS150.06E
CAUTION:
The nuts of the cassette edge sensor mounting can be lost very easily.
Only loosen the screws of the cassette edge sensor approx. 1 to 2 turns.
Always have a "small parts kit*" CM+9 5175 9070 1 with you: This contains 2 spare
screws and nuts for the cassette edge sensor mounting.
* The last digit in the spare part number indicates the spare part revision at release of this document.
When ordering, the actual revision of the spare part is delivered.
3
Figure 49
ASSEMBLY:
(1)
IMPORTANT:
Pin 2 is not connected at the cassette edge
sensor.
3
2 (not connected)
MECHANICAL ADJUSTMENT:
(1)
Edition 1, Revision 2
03-2009
CR 30-X
Type 5175 / 100/110
Chapter 3.5 / 22
DD+DIS150.06E
Verification:
Result:
2.3
2.4
Edition 1, Revision 2
03-2009
CR 30-X
Type 5175 / 100/110
Chapter 3.5 / 23
DD+DIS150.06E
2.5
Figure 50
REQUIRED TOOLS:
3 mm,
with ball-shaped head
REQUIRED TIME:
60 min.
DISASSEMBLY:
(1)
(2)
(3)
(4)
B
A
A
Figure 51
Edition 1, Revision 2
03-2009
CR 30-X
Type 5175 / 100/110
Chapter 3.5 / 24
DD+DIS150.06E
(5)
Figure 52
(6)
5175R3-5_004.cdr
Figure 53
(7)
(8)
Figure 54
(9)
Figure 55
DOCUMENT CONTROL NOTE:
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 2
03-2009
CR 30-X
Type 5175 / 100/110
Chapter 3.5 / 25
DD+DIS150.06E
Figure 56
(12) Take out cleaning brush holder.
Figure 57
(13) Remove screw at erasure unit.
(14) Disconnect two cables at erasure unit.
(15) Take out erasure unit.
Figure 58
CAUTION:
IP jams possible if IP guide plate is deformed.
Do not bend IP guide plate at removal.
Edition 1, Revision 2
03-2009
CR 30-X
Type 5175 / 100/110
Chapter 3.5 / 26
DD+DIS150.06E
Figure 59
(17) Disconnect cable at roller IP sensor.
Figure 60
(18) Open three cable holders at right hand
side.
Figure 61
(19) Open five cable holders at left hand side.
Figure 62
DOCUMENT CONTROL NOTE:
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 2
03-2009
CR 30-X
Type 5175 / 100/110
Chapter 3.5 / 27
DD+DIS150.06E
Figure 63
(21) Cut one cable tie at upper part transport
unit cable harness left hand side.
Figure 64
(22) Remove 2 screws of upper part transport
unit right hand side.
Figure 65
(23) Remove 2 screws of upper part transport
unit left hand side.
Figure 66
Edition 1, Revision 2
03-2009
CR 30-X
Type 5175 / 100/110
Chapter 3.5 / 28
DD+DIS150.06E
Figure 67
(25) Remove all cables at IP handling board.
(26) Open all cable holders at cassette unit.
(27) Remove IP handling board.
(28) Pull out cable for drawer motor.
Figure 68
(29) Turn a few turns at spindle of cassette
fixation (A).
(30) Remove two screws at the front (B, C).
Figure 69
(31) Remove 2 screws at cassette fixation unit.
Figure 70
DOCUMENT CONTROL NOTE:
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 2
03-2009
CR 30-X
Type 5175 / 100/110
Chapter 3.5 / 29
DD+DIS150.06E
Figure 71
(34) Remove cassette unlocking unit.
Figure 72
(35) Remove the cassette fixation unit.
Figure 73
Edition 1, Revision 2
03-2009
CR 30-X
Type 5175 / 100/110
Chapter 3.5 / 30
DD+DIS150.06E
ASSEMBLY:
(1)
CALIBRATION:
(1)
Verification:
(1)
(2)
(3)
(4)
(5)
(6)
Repeat steps (1) to (5) with each cassette format the customer is using.
IMPORTANT:
In case the boot-up is not successful, check following:
Cable is connected at user interface board.
Cable between chip reader and IP handling board is connected at P 18.
Result:
2.6
Figure 74
REQUIRED TOOLS:
3 mm,
with ball-shaped head
DOCUMENT CONTROL NOTE:
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 2
03-2009
CR 30-X
Type 5175 / 100/110
Chapter 3.5 / 31
DD+DIS150.06E
REQUIRED TIME:
60 min.
DISASSEMBLY:
(1)
(2)
(3)
(4)
B
A
A
Figure 75
(5)
Figure 76
(6)
5175R3-5_004.cdr
Figure 77
Edition 1, Revision 2
03-2009
CR 30-X
Type 5175 / 100/110
Chapter 3.5 / 32
DD+DIS150.06E
(7)
(8)
Figure 78
(9)
Figure 79
(10) Disconnect ground cable at right hand side.
(11) Remove front cover.
Figure 80
(12) Remove cable at P30 of IP handling board.
Figure 81
Edition 1, Revision 2
03-2009
CR 30-X
Type 5175 / 100/110
Chapter 3.5 / 33
DD+DIS150.06E
Figure 82
ASSEMBLY:
(1)
CALIBRATION:
(1)
Verification:
Result:
2.7
(1)
(2)
(3)
(4)
(5)
Figure 83
Edition 1, Revision 2
03-2009
CR 30-X
Type 5175 / 100/110
Chapter 3.5 / 34
DD+DIS150.06E
REQUIRED TOOLS:
3 mm
Small size
REQUIRED TIME:
60 min.
DISASSEMBLY:
(1)
(2)
(3)
(4)
B
A
A
Figure 84
(5)
Figure 85
Edition 1, Revision 2
03-2009
CR 30-X
Type 5175 / 100/110
Chapter 3.5 / 35
DD+DIS150.06E
(6)
5175R3-5_004.cdr
Figure 86
(7)
(8)
Figure 87
(9)
Figure 88
(10) Disconnect ground cable at right hand side.
(11) Remove front cover.
Figure 89
DOCUMENT CONTROL NOTE:
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 2
03-2009
CR 30-X
Type 5175 / 100/110
Chapter 3.5 / 36
DD+DIS150.06E
Figure 90
(15) Pull at the toothed belt to open the cassette
opener.
517536ak.cdr
Figure 91
(16) Disconnect cable at chip reader.
(17) Remove screw.
(18) Remove chip reader.
Figure 92
ASSEMBLY:
(1)
Edition 1, Revision 2
03-2009
CR 30-X
Type 5175 / 100/110
Chapter 3.5 / 37
DD+DIS150.06E
Verification:
Result:
2.8
(1)
(2)
(3)
(4)
(5)
(6)
(7)
2.9
Figure 93
REQUIRED TOOLS:
3 mm,
with ball-shaped head
DOCUMENT CONTROL NOTE:
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 2
03-2009
CR 30-X
Type 5175 / 100/110
Chapter 3.5 / 38
DD+DIS150.06E
REQUIRED TIME:
30 min.
DISASSEMBLY:
(1)
(2)
(3)
(4)
B
A
A
Figure 94
(5)
(6)
(7)
Figure 95
ASSEMBLY:
(1)
Edition 1, Revision 2
03-2009
CR 30-X
Type 5175 / 100/110
Chapter 3.5 / 39
DD+DIS150.06E
Verification:
Result:
2.10
(1)
(2)
(3)
(4)
(5)
(6)
(7)
Figure 96
REQUIRED TOOLS:
Medium size
REQUIRED TIME:
30 min.
Edition 1, Revision 2
03-2009
CR 30-X
Type 5175 / 100/110
Chapter 3.5 / 40
DD+DIS150.06E
DISASSEMBLY:
(1)
(2)
(3)
(4)
B
A
A
Figure 97
(5)
Figure 98
Figure 99
NOTE:
As the ball bearings are secured with Loctite, some force has to be applied to remove the
ball bearings.
For mounting of the ball bearings it is not required to secure them via Loctite again.
Edition 1, Revision 2
03-2009
CR 30-X
Type 5175 / 100/110
Chapter 3.5 / 41
DD+DIS150.06E
ASSEMBLY:
(1)
NOTE:
To support that the ball bearing fits in its
rail, lift the inner side plate slightly up.
Figure 100
Verification:
Result:
2.11
(1)
(2)
(3)
(4)
(5)
(6)
(7)
Edition 1, Revision 2
03-2009
CR 30-X
Type 5175 / 100/110
Chapter 3.5 / 42
DD+DIS150.06E
2.12
2.13
2.14
Figure 101
IMPORTANT:
Spare part CM+9 5175 6125 0 contains all 3 permanent magnets.
It is recommended to replace all 3 magnets together.
REQUIRED TOOLS:
3 mm,
with ball-shaped head
Medium size
Edition 1, Revision 2
03-2009
CR 30-X
Type 5175 / 100/110
Chapter 3.5 / 43
DD+DIS150.06E
REQUIRED TIME:
30 min.
DISASSEMBLY:
(1)
(2)
(3)
(4)
B
A
A
Figure 102
(5)
Figure 103
(6)
5175R3-5_004.cdr
Figure 104
Edition 1, Revision 2
03-2009
CR 30-X
Type 5175 / 100/110
Chapter 3.5 / 44
DD+DIS150.06E
(7)
(8)
Figure 105
(9)
Figure 106
(10) Disconnect ground cable at right hand side.
(11) Remove front cover.
Figure 107
(12) Remove screw at erasure unit.
(13) Disconnect two cables at erasure unit.
(14) Take out erasure unit.
Figure 108
Edition 1, Revision 2
03-2009
CR 30-X
Type 5175 / 100/110
Chapter 3.5 / 45
DD+DIS150.06E
CAUTION:
IP jams possible if IP guide plate is deformed.
Do not bend IP guide plate at removal.
(15) Slide IP guide plate to the left and lift it out.
Figure 109
(16) Remove screws at magnets.
(17) Remove magnets.
517535ai.cdr
Figure 110
ASSEMBLY:
(1)
Edition 1, Revision 2
03-2009
CR 30-X
Type 5175 / 100/110
Chapter 3.5 / 46
DD+DIS150.06E
Verification:
Result:
2.15
(1)
(2)
(3)
(4)
(5)
(6)
(7)
2.16
Figure 111
Edition 1, Revision 2
03-2009
CR 30-X
Type 5175 / 100/110
Chapter 3.5 / 47
DD+DIS150.06E
REQUIRED TOOLS:
3 mm,
with ball-shaped head
REQUIRED TIME:
30 min.
DISASSEMBLY:
(1)
(2)
(3)
(4)
B
A
A
Figure 112
(5)
Figure 113
Edition 1, Revision 2
03-2009
CR 30-X
Type 5175 / 100/110
Chapter 3.5 / 48
DD+DIS150.06E
(6)
5175R3-5_004.cdr
Figure 114
(7)
(8)
Figure 115
(9)
Figure 116
(10) Disconnect ground cable at right hand side.
(11) Remove front cover.
Figure 117
DOCUMENT CONTROL NOTE:
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 2
03-2009
CR 30-X
Type 5175 / 100/110
Chapter 3.5 / 49
DD+DIS150.06E
Figure 118
(14) Open 3 screws at the fixer stepper motor
(only 2 screws are shown in Figure 119).
(15) Remove the fixer stepper motor.
Figure 119
ASSEMBLY:
(1)
IMPORTANT:
Mount the motor with the motor cable outlet as
shown in the figure enclosed:
Figure 120
Edition 1, Revision 2
03-2009
CR 30-X
Type 5175 / 100/110
Chapter 3.5 / 50
DD+DIS150.06E
CALIBRATION:
(1)
Verification:
2.17
(1)
(2)
(3)
(4)
(5)
Figure 121
REQUIRED TOOLS:
3 mm,
with ball-shaped head
REQUIRED TIME:
30 min.
Edition 1, Revision 2
03-2009
CR 30-X
Type 5175 / 100/110
Chapter 3.5 / 51
DD+DIS150.06E
DISASSEMBLY:
(1)
(2)
(3)
(4)
B
A
A
Figure 122
(5)
Figure 123
(6)
5175R3-5_004.cdr
Figure 124
Edition 1, Revision 2
03-2009
CR 30-X
Type 5175 / 100/110
Chapter 3.5 / 52
DD+DIS150.06E
(7)
(8)
Figure 125
(9)
Figure 126
(10) Disconnect ground cable at right hand side.
(11) Remove front cover.
Figure 127
(12) Remove cable at plug P19 of IP handling
board.
(13) Open cable holders.
Figure 128
Edition 1, Revision 2
03-2009
CR 30-X
Type 5175 / 100/110
Chapter 3.5 / 53
DD+DIS150.06E
Figure 129
ASSEMBLY:
(1)
IMPORTANT:
Mount the motor with the motor cable outlet as
shown in the figure enclosed:
Figure 130
CALIBRATION:
(1)
Edition 1, Revision 2
03-2009
CR 30-X
Type 5175 / 100/110
Chapter 3.5 / 54
DD+DIS150.06E
Verification
2.18
(1)
(2)
(3)
(4)
(5)
Figure 131
REQUIRED TOOLS:
3 mm,
with ball-shaped head
REQUIRED TIME:
30 min.
Edition 1, Revision 2
03-2009
CR 30-X
Type 5175 / 100/110
Chapter 3.5 / 55
DD+DIS150.06E
DISASSEMBLY:
(1)
(2)
(3)
(4)
B
A
A
Figure 132
(5)
Figure 133
(6)
5175R3-5_004.cdr
Figure 134
Edition 1, Revision 2
03-2009
CR 30-X
Type 5175 / 100/110
Chapter 3.5 / 56
DD+DIS150.06E
(7)
(8)
Figure 135
(9)
Figure 136
(10) Disconnect ground cable at right hand side.
(11) Remove front cover.
Figure 137
(16) Remove cable at plug P25 of IP handling
board.
(17) Open cable holders.
Figure 138
DOCUMENT CONTROL NOTE:
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 2
03-2009
CR 30-X
Type 5175 / 100/110
Chapter 3.5 / 57
DD+DIS150.06E
Figure 139
ASSEMBLY:
(1)
Figure 140
Verification
(1)
(2)
(3)
(4)
(5)
(6)
(7)
Edition 1, Revision 2
03-2009
CR 30-X
Type 5175 / 100/110
Chapter 3.5 / 58
DD+DIS150.06E
2.19
Figure 141
REQUIRED TOOLS:
3 mm,
with ball-shaped head
REQUIRED TIME:
30 min.
DISASSEMBLY:
(1)
(2)
(3)
(4)
B
A
A
Figure 142
Edition 1, Revision 2
03-2009
CR 30-X
Type 5175 / 100/110
Chapter 3.5 / 59
DD+DIS150.06E
(5)
Figure 143
(6)
5175R3-5_004.cdr
Figure 144
(7)
(8)
Figure 145
(9)
Figure 146
DOCUMENT CONTROL NOTE:
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 2
03-2009
CR 30-X
Type 5175 / 100/110
Chapter 3.5 / 60
DD+DIS150.06E
Figure 147
(12) Take out cleaning brush holder.
Figure 148
(13) Disconnect two cables at erasure unit.
(14) Remove screw at erasure unit.
(15) Take out erasure unit.
Figure 149
CAUTION:
IP jams possible if IP guide plate is deformed.
Do not bend IP guide plate at removal.
Edition 1, Revision 2
03-2009
CR 30-X
Type 5175 / 100/110
Chapter 3.5 / 61
DD+DIS150.06E
Figure 150
(17) Remove cable at plug P29 of IP handling
board.
(18) Open cable holders.
Figure 151
(19) Remove 4 screws of stepper motor roller
clamping (only 3 screws are shown in
Figure 152).
(20) Remove motor.
Figure 152
ASSEMBLY:
(1)
Edition 1, Revision 2
03-2009
CR 30-X
Type 5175 / 100/110
Chapter 3.5 / 62
DD+DIS150.06E
IMPORTANT:
Mount the motor with the motor cable outlet as
shown in the figure enclosed (see arrow):
Figure 153
CALIBRATION:
(1)
Verification
(1)
(2)
(3)
(4)
(5)
Edition 1, Revision 2
03-2009
CR 30-X
Type 5175 / 100/110
Chapter 3.5 / 63
DD+DIS150.06E
2.20
Figure 154
REQUIRED TOOLS:
None.
REQUIRED TIME:
15 min.
DISASSEMBLY:
(1)
(2)
(3)
(4)
B
A
A
Figure 155
Edition 1, Revision 2
03-2009
CR 30-X
Type 5175 / 100/110
Chapter 3.5 / 64
DD+DIS150.06E
(5)
(6)
Figure 156
ASSEMBLY:
(1)
Verification:
Result:
(1)
(2)
(3)
(4)
(5)
(6)
(7)
Edition 1, Revision 2
03-2009
CR 30-X
Type 5175 / 100/110
Chapter 3.5 / 65
DD+DIS150.06E
2.21
Figure 157
IMPORTANT:
Spare part CM+9 5175 9050 0 contains all replaceable belts for the digitizer.
It is not required to replace all toothed belts together.
Only replace the defective toothed belt.
REQUIRED TOOLS:
3 mm,
with ball-shaped head
Small size
REQUIRED TIME:
90 min.
DISASSEMBLY:
(1)
(2)
(3)
(4)
B
A
A
Figure 158
DOCUMENT CONTROL NOTE:
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 2
03-2009
CR 30-X
Type 5175 / 100/110
Chapter 3.5 / 66
DD+DIS150.06E
(5)
Figure 159
(6)
5175R3-5_004.cdr
Figure 160
(7)
(8)
Figure 161
(9)
Figure 162
DOCUMENT CONTROL NOTE:
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 2
03-2009
CR 30-X
Type 5175 / 100/110
Chapter 3.5 / 67
DD+DIS150.06E
Figure 163
(12) Take out cleaning brush holder.
Figure 164
(13) Disconnect two cables at erasure unit.
(14) Remove screw at erasure unit.
(15) Take out erasure unit.
Figure 165
CAUTION:
IP jams possible if IP guide plate is deformed.
Do not bend IP guide plate at removal.
Edition 1, Revision 2
03-2009
CR 30-X
Type 5175 / 100/110
Chapter 3.5 / 68
DD+DIS150.06E
Figure 166
(17) Disconnect cable at roller IP sensor.
Figure 167
(18) Open three cable holders at right hand
side.
Figure 168
(19) Open cable holders at left hand side.
Figure 169
DOCUMENT CONTROL NOTE:
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 2
03-2009
CR 30-X
Type 5175 / 100/110
Chapter 3.5 / 69
DD+DIS150.06E
Figure 170
(21) Cut one cable tie at upper part transport
unit cable harness left hand side.
Figure 171
(22) Unplug P4, P17, P25 and P29 at IP
handling board.
Figure 172
(23) Remove 2 screws of upper part transport
unit right hand side.
Figure 173
Edition 1, Revision 2
03-2009
CR 30-X
Type 5175 / 100/110
Chapter 3.5 / 70
DD+DIS150.06E
Figure 174
(25) Take off upper part transport unit.
Figure 175
(26) With a screwdriver remove two ball
bearings at transport unit left hand side.
Figure 176
(27) Remove two ball bearings at transport unit
right hand side.
Figure 177
Edition 1, Revision 2
03-2009
CR 30-X
Type 5175 / 100/110
Chapter 3.5 / 71
DD+DIS150.06E
Figure 178
(29) Open screws at axle gear plate.
Figure 179
(30) Remove c-clip at axle.
Figure 180
Edition 1, Revision 2
03-2009
CR 30-X
Type 5175 / 100/110
Chapter 3.5 / 72
DD+DIS150.06E
Figure 181
ASSEMBLY:
(1)
CALIBRATION:
(1)
Verification:
Result:
(1)
(2)
(3)
(4)
(5)
(6)
Repeat steps (1) to (5) with each cassette format the customer is using.
Edition 1, Revision 2
03-2009
CR 30-X
Type 5175 / 100/110
Chapter 3.5 / 73
DD+DIS150.06E
2.22
Figure 182
IMPORTANT:
Spare part CM+9 5175 9050 0 contains all belts for the digitizer.
It is not required to replace all toothed belts together.
Only replace the defective toothed belt.
REQUIRED TOOLS:
3 mm,
with ball-shaped head
Small size
REQUIRED TIME:
30 min.
DISASSEMBLY:
(1)
(2)
(3)
(4)
B
A
A
Figure 183
Edition 1, Revision 2
03-2009
CR 30-X
Type 5175 / 100/110
Chapter 3.5 / 74
DD+DIS150.06E
(5)
Figure 184
(6)
5175R3-5_004.cdr
Figure 185
(7)
(8)
Figure 186
(9)
Figure 187
DOCUMENT CONTROL NOTE:
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 2
03-2009
CR 30-X
Type 5175 / 100/110
Chapter 3.5 / 75
DD+DIS150.06E
Figure 188
(12) Remove screw at erasure unit.
(13) Disconnect two cables at erasure unit.
(14) Take out erasure unit.
Figure 189
CAUTION:
IP jams possible if IP guide plate is deformed.
Do not bend IP guide plate at removal.
(15) Remove IP guide plate.
A
B
Figure 190
Edition 1, Revision 2
03-2009
CR 30-X
Type 5175 / 100/110
Chapter 3.5 / 76
DD+DIS150.06E
517535aj.cdr
Figure 191
ASSEMBLY:
(1)
CALIBRATION:
(1)
Verification:
Result:
2.23
(1)
(2)
(3)
(4)
(5)
Figure 192
DOCUMENT CONTROL NOTE:
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 2
03-2009
CR 30-X
Type 5175 / 100/110
Chapter 3.5 / 77
DD+DIS150.06E
IMPORTANT:
Spare part CM+9 5175 9050 0 contains all belts for the digitizer.
It is not required to replace all toothed belts together.
Only replace the defective toothed belt.
REQUIRED TOOLS:
3 mm
Small size
REQUIRED TIME:
30 min.
DISASSEMBLY:
(1)
(2)
(3)
(4)
B
A
A
Figure 193
(5)
Figure 194
Edition 1, Revision 2
03-2009
CR 30-X
Type 5175 / 100/110
Chapter 3.5 / 78
DD+DIS150.06E
(6)
5175R3-5_004.cdr
Figure 195
(7)
(8)
Figure 196
(9)
Figure 197
(10) Disconnect ground cable at right hand side.
(11) Remove front cover.
Figure 198
DOCUMENT CONTROL NOTE:
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 2
03-2009
CR 30-X
Type 5175 / 100/110
Chapter 3.5 / 79
DD+DIS150.06E
Figure 199
ASSEMBLY:
(1)
Verification:
Result:
(1)
(2)
(3)
(4)
(5)
(6)
(7)
Edition 1, Revision 2
03-2009
CR 30-X
Type 5175 / 100/110
Chapter 3.5 / 80
DD+DIS150.06E
2.24
Figure 200
IMPORTANT:
Spare part CM+9 5175 9050 0 contains all belts for the digitizer.
It is not required to replace all toothed belts together.
Only replace the defective toothed belt.
REQUIRED TOOLS:
3 mm
Small size
REQUIRED TIME:
30 min.
DISASSEMBLY:
(1)
(2)
(3)
(4)
B
A
A
Figure 201
Edition 1, Revision 2
03-2009
CR 30-X
Type 5175 / 100/110
Chapter 3.5 / 81
DD+DIS150.06E
(5)
Figure 202
(6)
5175R3-5_004.cdr
Figure 203
(7)
(8)
Figure 204
(9)
Figure 205
DOCUMENT CONTROL NOTE:
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 2
03-2009
CR 30-X
Type 5175 / 100/110
Chapter 3.5 / 82
DD+DIS150.06E
Figure 206
(12) Remove toothed belt.
Figure 207
ASSEMBLY:
(1)
CALIBRATION:
(1)
Verification:
Result:
(1)
(2)
(3)
(4)
(5)
Edition 1, Revision 2
03-2009
CR 30-X
Type 5175 / 100/110
Chapter 3.5 / 83
DD+DIS150.06E
2.25
Figure 208
REQUIRED TOOLS:
3 mm
Medium size
REQUIRED TIME:
5 min.
DISASSEMBLY:
(1)
(2)
(3)
(4)
B
A
A
Figure 209
Edition 1, Revision 2
03-2009
CR 30-X
Type 5175 / 100/110
Chapter 3.5 / 84
DD+DIS150.06E
(5)
Figure 210
(6)
5175R3-5_004.cdr
Figure 211
(7)
(8)
Figure 212
(9)
Figure 213
DOCUMENT CONTROL NOTE:
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 2
03-2009
CR 30-X
Type 5175 / 100/110
Chapter 3.5 / 85
DD+DIS150.06E
Figure 214
(12) Remove four screws.
(13) Remove the user interface board.
Figure 215
ASSEMBLY:
(1)
Verification:
Result:
(1)
(2)
(3)
(4)
(5)
(6)
(7)
Edition 1, Revision 2
03-2009
CR 30-X
Type 5175 / 100/110
Chapter 3.5 / 86
DD+DIS150.06E
3.1
3.2
3.3
Figure 216
Edition 1, Revision 2
03-2009
CR 30-X
Type 5175 / 100/110
Chapter 3.5 / 87
DD+DIS150.06E
REQUIRED TOOLS:
Antistatic wrist strap
3 mm
5,5 mm (magnetic)
REQUIRED TIME:
30 min.
DISASSEMBLY:
(1)
(2)
(3)
(4)
B
A
A
Figure 217
(5)
(6)
Figure 218
Edition 1, Revision 2
03-2009
CR 30-X
Type 5175 / 100/110
Chapter 3.5 / 88
DD+DIS150.06E
(7)
(8)
(9)
5175R3-3_103
Figure 219
ASSEMBLY:
(1)
Verification:
Result:
3.4
Edition 1, Revision 2
03-2009
CR 30-X
Type 5175 / 100/110
Chapter 3.5 / 89
DD+DIS150.06E
3.5
3.6
Figure 220
IMPORTANT:
Spare part CM+9 5175 2650 1 contains both upper scan rollers. It is recommended to
replace both scan rollers together.
REQUIRED TOOLS:
5 mm
3 mm
REQUIRED TIME:
45 min.
Edition 1, Revision 2
03-2009
CR 30-X
Type 5175 / 100/110
Chapter 3.5 / 90
DD+DIS150.06E
DISASSEMBLY:
(1)
(2)
(3)
(4)
B
A
A
Figure 221
(5)
(6)
Figure 222
(7)
(8)
Figure 223
Edition 1, Revision 2
03-2009
CR 30-X
Type 5175 / 100/110
Chapter 3.5 / 91
DD+DIS150.06E
(9)
Figure 224
(11) Remove screws of the rear cover.
(12) Remove rear cover.
Figure 225
(13) Remove safety bracket at front scan roller.
(14) Remove front scan roller.
Figure 226
(15) Remove safety bracket at rear scan roller
left hand side.
Figure 227
Edition 1, Revision 2
03-2009
CR 30-X
Type 5175 / 100/110
Chapter 3.5 / 92
DD+DIS150.06E
Figure 228
ASSEMBLY:
(1) Mount in reverse order.
Verification:
Result:
3.7
Edition 1, Revision 2
03-2009
CR 30-X
Type 5175 / 100/110
Chapter 3.5 / 93
DD+DIS150.06E
3.8
3.9
4.1
Figure 229
REQUIRED TOOLS:
3 mm
5 mm
REQUIRED TIME:
60 min.
Edition 1, Revision 2
03-2009
CR 30-X
Type 5175 / 100/110
Chapter 3.5 / 94
DD+DIS150.06E
DISASSEMBLY:
(1)
(2)
(3)
(4)
B
A
A
Figure 230
(5)
(6)
Figure 231
(7)
(8)
Figure 232
Edition 1, Revision 2
03-2009
CR 30-X
Type 5175 / 100/110
Chapter 3.5 / 95
DD+DIS150.06E
(9)
Figure 233
ASSEMBLY:
(1)
MECHANICAL ADJUSTMENT:
Verification:
Result:
(1)
(1)
(2)
(3)
(4)
(5)
(6)
(7)
Edition 1, Revision 2
03-2009
CR 30-X
Type 5175 / 100/110
Chapter 3.5 / 96
DD+DIS150.06E
4.2
Figure 234
REQUIRED TOOLS:
3 mm
REQUIRED TIME:
30 min.
DISASSEMBLY:
(1)
(2)
(3)
(4)
B
A
A
Figure 235
Edition 1, Revision 2
03-2009
CR 30-X
Type 5175 / 100/110
Chapter 3.5 / 97
DD+DIS150.06E
(5)
(6)
Figure 236
(7)
(8)
Figure 237
(9)
Figure 238
(11) Remove screws of the rear cover.
(12) Remove rear cover.
Figure 239
DOCUMENT CONTROL NOTE:
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 2
03-2009
CR 30-X
Type 5175 / 100/110
Chapter 3.5 / 98
DD+DIS150.06E
Figure 240
ASSEMBLY:
(1) Mount in reverse order.
Verification:
Result:
(1)
(2)
(3)
(4)
(5)
(6)
(7)
Edition 1, Revision 2
03-2009
CR 30-X
Type 5175 / 100/110
Chapter 3.5 / 99
DD+DIS150.06E
4.3
Figure 241
REQUIRED TOOLS:
3 mm
Medium size
5 mm
2,5 mm
REQUIRED TIME:
60 min.
DANGER:
If the power cord is connected and the main switch is switched on the input
side of the power supply is connected to high voltage. Risk of live.
Remove all cables in the rear of the digitizer before removal of the rear cover
Edition 1, Revision 2
03-2009
CR 30-X
Type 5175 / 100/110
DD+DIS150.06E
DISASSEMBLY:
(1)
(2)
(3)
(4)
B
A
A
Figure 242
(5)
(6)
Figure 243
(7)
(8)
Figure 244
Edition 1, Revision 2
03-2009
CR 30-X
Type 5175 / 100/110
DD+DIS150.06E
(9)
Figure 245
(11) Remove screws of the rear cover.
(12) Remove rear cover.
Figure 246
(13) Remove 2 rear screws.
Figure 247
(14) Unscrew all connectors at output side of
power supply.
Figure 248
DOCUMENT CONTROL NOTE:
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 2
03-2009
CR 30-X
Type 5175 / 100/110
DD+DIS150.06E
Figure 249
DANGER:
If the cables at the input side of the power
supply are mounted wrong, the housing of
the digitizer can be connected to high
voltage. Risk of live.
Mount the cables at the power supply input as
indicated on the cables and shown in Figure
250.
Figure 250
ASSEMBLY:
(1) Mount in reverse order.
See IMPORTANT note below.
black red
IMPORTANT:
Connect the cables at the output side as shown
in Figure 251.
Figure 251
DOCUMENT CONTROL NOTE:
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 2
03-2009
CR 30-X
Type 5175 / 100/110
DD+DIS150.06E
4.3.1
200
517511ai.cdr
Figure 252
(4)
(5)
(6)
517509ah.cdr
Figure 253
IMPORTANT:
In case the electrical resistance of one measurement exceeds the limits (0.2 Ohm
without mains cable; 0.3 Ohm with mains cable) check the digitizer for loose grounding
connections!
Edition 1, Revision 2
03-2009
CR 30-X
Type 5175 / 100/110
DD+DIS150.06E
4.3.2
4.3.3
Verification:
Result:
Edition 1, Revision 2
03-2009
CR 30-X
Type 5175 / 100/110
Chapter 3.6
HEALTHCARE
Imaging Services
CR 30-X
Type 5175 / 100/110
Document History
Edition.
Revision
Release
Date
Changes
compared to previous Version 1.2:
1.3
10-2009
Referenced Documents
Document
Title
DD+DIS150.06E
Edition 1, Revision 3
10-2009 printed in Germany
DD+DIS150.06E
Manufacturer
Agfa HealthCare N.V.
Publisher
Agfa-Gevaert HealthCare GmbH
Tegernseer Landstrae 161
D - 81539 Mnchen
Germany
Copyright 2009 Agfa HealthCare N.V.
All rights reserved.
Technical modifications reserved.
AGFA and the Agfa-Rhombus are trademarks of Agfa HealthCare N.V.
WARNING:
Improper operation or service activities may cause damage or injuries.
INSTRUCTION:
(1)
(2)
Strictly observe all safety directions within the "Generic Safety Directions" and on
the product.
NOTE:
To verify the latest version of single documents and of Service Manuals refer to the
Document Type Order List in the GSO library.
Edition 1, Revision 3
10-2009
CR 30-X
Type 5175 / 100/110
Chapter 3.6 / 2
DD+DIS150.06E
LIST OF CONTENTS
ADJUSTMENTS .......................................................................................................................5
2.1
2.1.1
2.1.2
2.1.3
2.2
2.3
2.3.1
2.3.2
2.4
2.5
2.5.1
2.5.2
2.5.3
CALIBRATIONS .....................................................................................................................29
3.1
Shading Calibration.................................................................................................................29
3.1.1
3.1.2
3.1.3
Appendix: Changes in the Shading Calibration Procedure if Full Leg / Full Spine Cassettes
are in Use................................................................................................................................33
3.2
Edition 1, Revision 3
10-2009
CR 30-X
Type 5175 / 100/110
Chapter 3.6 / 3
DD+DIS150.06E
Adjustment Type
Use Case
Section
Drawer Unit
Adjustment
2.1
2.2
Cassette unit
safety switch
position adjustment
2.3
Cassette edge
sensor position
adjustment
2.4
Laser beam
position check and
adjustment
Calibration Type
Shading calibration
Stall Calibration
2.5
Use Case
Section
3.1
3.2
Edition 1, Revision 3
10-2009
CR 30-X
Type 5175 / 100/110
Chapter 3.6 / 4
DD+DIS150.06E
Adjustments
2.1
Purpose
Use Case
Adjustment of the optimum height and lateral position of the upper drawer unit plate
for reliable image plate transport in and out of the cassette.
NOTE:
Two versions of the drawer adjustment tool are available:
max.
Version 1, delivered
till October 2009
min.
Version 2, delivered
as of October 2009
Figure 1
Drawer Unit
Drawer Motor
Its is also available from spare parts stock with spare part number*: CM+9 5175 6001 0
* The last digit in the spare part number indicates the spare part revision at release of
this document. When ordering, the actual revision of the spare part is delivered.
This document describes the adjustment with drawer adjustment tool version 2.
Where applicable, differences to version 1 are described.
REQUIRED TIME:
30 minutes
Edition 1, Revision 3
10-2009
CR 30-X
Type 5175 / 100/110
Chapter 3.6 / 5
DD+DIS150.06E
2.1.1
Section
2.1.1
2.1.2
2.1.3
(2)
(3)
(4)
B
A
A
Figure 2
CAUTION:
Less reliable IP run possible if IP guide plate is deformed.
Do not bend IP guide plate at removal.
(5)
(6)
(7)
Figure 3
Edition 1, Revision 3
10-2009
CR 30-X
Type 5175 / 100/110
Chapter 3.6 / 6
DD+DIS150.06E
(8)
517536ak.cdr
Figure 4
CAUTION:
Misalignment of drawer unit possible when pushing it at the upper plate.
When pushing the drawer unit only do this via drawer unit axle.
(9)
Upper plate
Drawer axle
Figure 5
2.1.2
(2)
(3)
Edition 1, Revision 3
10-2009
CR 30-X
Type 5175 / 100/110
Chapter 3.6 / 7
DD+DIS150.06E
NOT OK
NOT OK
OK
Lateral stop
Notch
Distance exaggerated
Distance exaggerated
517536ah.cdr
Figure 7
IMPORTANT:
When inserting the drawer unit adjustment tool, be sure, that the notch of the tool is
properly inserted in the right hand side of the cassette unit entry. See Figure 7.
Evaluation
In case the lateral stop is aligned with the plastic finger at the drawer unit there is no
adjustment required. Continue with section 2.1.3.
Otherwise perform lateral adjustment as described in next steps.
LATERAL ADJUSTMENT
(1)
517536dd.cdr
Figure 8
Edition 1, Revision 3
10-2009
CR 30-X
Type 5175 / 100/110
Chapter 3.6 / 8
DD+DIS150.06E
(2)
(3)
2.1.3
Plastic finger
Cassette input
side wall
Lateral stop
517536az.cdr
Figure 9
Insert the drawer unit adjustment tool in the cassette unit entry.
(2)
Check the minimum and maximum height. See also IMPORTANT note below.
Check min.
height right.
Check min.
height left.
min.
max.
max.
min.
Check max.
height right.
Check max.
height left.
max.
max.
min.
min.
517536an.cdr
Figure 10
IMPORTANT:
Perform the min. height check between Magnet
magnet and plastic finger.
Plastic Finger
min.
Figure 11
Edition 1, Revision 3
10-2009
CR 30-X
Type 5175 / 100/110
Chapter 3.6 / 9
DD+DIS150.06E
Expected result:
Min. height
Max. height
Figure 12
NOTE:
With drawer adjustment tool version 1 it is possible to check the minimum height only.
It is not possible to check the maximum height.
Evaluation
HEIGHT ADJUSTMENT
(1)
517536cd.cdr
Figure 13
Edition 1, Revision 3
10-2009
CR 30-X
Type 5175 / 100/110
Chapter 3.6 / 10
DD+DIS150.06E
Verification
Result
(2)
(3)
(4)
(5)
(6)
(7)
(8)
(1)
(2)
(3)
Scan one image plate with each cassette format the customer is using.
Figure 14
Edition 1, Revision 3
10-2009
CR 30-X
Type 5175 / 100/110
Chapter 3.6 / 11
DD+DIS150.06E
2.2
Purpose
Use Case
13 mm
6 mm
PMT gauge
(Part of delivery)
REQUIRED TIME:
30 minutes
IMPORTANT:
The PMI Board Carrier Position Adjustment is part of the replacement procedure PMT
with Light Collector.
The adjustment has to be done before mounting the PMT, after mounting optic module
and PMI board carrier.
MECHANICAL ADJUSTMENT:
(1)
(2)
Figure 15
Edition 1, Revision 3
10-2009
CR 30-X
Type 5175 / 100/110
Chapter 3.6 / 12
DD+DIS150.06E
(3)
Figure 16
(4)
(5)
(6)
(7)
(8)
(9)
Figure 17
Figure 18
Edition 1, Revision 3
10-2009
CR 30-X
Type 5175 / 100/110
Chapter 3.6 / 13
DD+DIS150.06E
Result
2.3
Purpose
Use Case
The PMI board carrier is adjusted this way, that the light collector is mounted
stress free.
3 mm
REQUIRED TIME:
20 minutes
Edition 1, Revision 3
10-2009
CR 30-X
Type 5175 / 100/110
Chapter 3.6 / 14
DD+DIS150.06E
2.3.1
(2)
(3)
(4)
(5)
1 cm
Figure 20
2.3.2
(2)
(3)
(4)
(5)
(6)
(7)
(8)
Mount covers.
Figure 21
Edition 1, Revision 3
10-2009
CR 30-X
Type 5175 / 100/110
Chapter 3.6 / 15
DD+DIS150.06E
Verification
Result
2.4
Purpose
Use Case
(1)
(2)
(3)
(4)
Confirm that the switch opens and switches off the digitizer.
(5)
(6)
REQUIRED TOOLS:
Optional:
1,5 mm
200
Multimeter
(Ohmmeter)
REQUIRED TIME:
20 minutes
Edition 1, Revision 3
10-2009
CR 30-X
Type 5175 / 100/110
Chapter 3.6 / 16
DD+DIS150.06E
(1)
(2)
(3)
(4)
(5)
B
A
A
Figure 22
MECHANICAL ADJUSTMENT:
(1)
(2)
(3)
(4)
Fixation
screws
Right:
Black tube side on top
Figure 23
Edition 1, Revision 3
10-2009
CR 30-X
Type 5175 / 100/110
Chapter 3.6 / 17
DD+DIS150.06E
Verification
Result
(5)
(6)
(7)
(8)
Fixation
screws
Wrong:
Red side on top
Figure 24
(1)
(2)
(3)
Insert a cassette the wrong way: The digitizer may not recognize the cassette.
(4)
Insert the cassette the right way: The digitizer has to recognize the cassette.
Edition 1, Revision 3
10-2009
CR 30-X
Type 5175 / 100/110
Chapter 3.6 / 18
DD+DIS150.06E
2.5
Purpose
Use Case
Checking the laser beam position, to see, whether it is centered to the calibration
board.
Adjustment of the optic module this way that the laser beam position is centered
to the calibration board.
Perform the "Laser Beam Position Check and Adjustment" in following cases:
REQUIRED TOOLS:
3 mm
REQUIRED TIME:
Approximately 30 minutes
IMPORTANT:
This adjustment requires:
Software CRE_1511**
For digitizers with lower serial numbers, which do not fulfil the requirements, it can also
be performed with the Laser Beam Detection Toolkit, spare part number***:
CM+9 5175 2940 0
The Laser Beam Detection Toolkit has instructions for installation and adjustment
included.
* Installed in SN 6397 (Type 5175/100) and SN 5935 (Type 5175/110) and in all digitizers, where the
calibration board was exchanged after September 2009.
** Installed in SN 5570 (Type 5175/100) and SN 6377 (Type 5175/110).
*** The last digit in the spare part number indicates the spare part revision at release of
this document. When ordering, the actual revision of the spare part is delivered.
Edition 1, Revision 3
10-2009
CR 30-X
Type 5175 / 100/110
Chapter 3.6 / 19
DD+DIS150.06E
Distributor Board
Version 1
Distributor Board
Version 2
5175_enc900.cdr
The laser beam position check and adjustment is divided in following steps:
#
Step
Section
2.5.1
2.5.2
2.5.3
Edition 1, Revision 3
10-2009
CR 30-X
Type 5175 / 100/110
Chapter 3.6 / 20
DD+DIS150.06E
NOTE:
Short description of the laser beam position check:
By putting the switch on the distributor board to ON position the laser beam
position check functionality is activated: The yellow "LED 1" lights up.
If "LED 2" and "LED 3" on the distribution board are light up, the laser beam
position is OK.
By putting the switch on the distributor board to ON position the laser beam
position check functionality is activated: The yellow "LED 1" lights up.
By opening the screws at the two Optic Module Position Plates the Optic Module
can be shifted this way that both Laser Beam Position Detection Sensors are
covered by the laser beam. In this case LED 2 and LED 3 are light up.
Optic Module
Position Plate 2
Optic Module
Optic Module
Position Plate 1
EOL Sensor
Laser Beam Position
Detection Sensor 1
LED 2
LED 3
Distributor Board
Version 2
Switch
LED 1
BOL Sensor
517502by.cdr
Calibration Board
Version 2
Figure 26: Functional principle of the laser beam position check and adjustment
Edition 1, Revision 3
10-2009
CR 30-X
Type 5175 / 100/110
Chapter 3.6 / 21
DD+DIS150.06E
2.5.1
(2)
(3)
(4)
(5)
B
A
A
Figure 27
(6)
517536df.cdr
OFF
ON
Figure 28
DANGER:
If the power cord is connected and the main switch is switched on the input
side of the power supply is connected to high voltage. Risk of live.
Do not touch the input side of the power supply by hand or with any tool.
(7)
(8)
(9)
LED 1
Figure 29
Edition 1, Revision 3
10-2009
CR 30-X
Type 5175 / 100/110
Chapter 3.6 / 22
DD+DIS150.06E
Figure 30
WARNING:
Laser light. Eye injuries possible.
Do not keep any tool in the laser beam while the laser is switched on.
(15) Select the button <On> to switch on the laser beam. See Figure 31.
The status message shows <Polygon is spinning. Laser ready>.
See Figure 32. The Polygon motor is audible.
Figure 31
Figure 32
Edition 1, Revision 3
10-2009
CR 30-X
Type 5175 / 100/110
Chapter 3.6 / 23
DD+DIS150.06E
NOTE:
The laser beam switches off automatically after 30 minutes.
(16) Check the LEDs on the distribution board.
LED 2
LED 3
LED 1
Switch
Figure 33
Status
Action
LED 1 ON
LED 2 ON
LED 3 ON
LED 1 ON
LED 2 OFF
LED 3 ON
LED 1 ON
LED 2 ON
LED 3 OFF
LED 1 ON
LED 2 OFF
LED 3 OFF
LED 1
OFF
LED 2 OFF
LED 3 OFF
Edition 1, Revision 3
10-2009
CR 30-X
Type 5175 / 100/110
Chapter 3.6 / 24
DD+DIS150.06E
CAUTION:
If switch on distributor board stays in "ON" position during digitizer operation,
reduced image quality possible.
Set the switch on the distributor board to the OFF position before handing the digitizer
over to the customer.
(18) Set the switch on the distributor
board to the OFF position.
(19) Mount all covers.
(20) Connect the cables in the rear.
(21) Continue with section 2.5.3.
OFF
ON
Figure 34
Result
2.5.2
Figure 35
Edition 1, Revision 3
10-2009
CR 30-X
Type 5175 / 100/110
Chapter 3.6 / 25
DD+DIS150.06E
(2)
Figure 36
(3)
LED 3
Switch
LED 2
LED 1
Figure 37
(4)
Figure 38
Edition 1, Revision 3
10-2009
CR 30-X
Type 5175 / 100/110
Chapter 3.6 / 26
DD+DIS150.06E
(5)
(6)
Figure 39
CAUTION:
If switch on distributor board stays in "ON" position during digitizer operation,
reduced image quality possible.
Set the switch on the distributor board to the OFF position before handing the digitizer
over to the customer.
(7)
(8)
(9)
This device complies with Part 15 of the FCC rules. Operation is subject to the following two conditions:
(1) This device may not cuase harmful interference (2) This device must accept any interference
received, including interference that may cause undesired operation.
This Class A digital apparatus meets all requirements of the Canadian Interference Causing Equipment
Regulations.
Cet appareil numerique de la classe A respecte toutes les exigences du Reglement sur le material
brouillant du Canada.
Type B equipment
Gerte Typ B
Appareil du type B
ME D IC A L EL E CT R IC A L
EQU IP ME NT
WI TH RE SP E CT T O
E L EC TR I C S HOC K , F R
I E,
A ND MEC H A NI C AL
H AZ A RD S ON LY I N
AC C ORD A N C E WI TH
U L6 060 1-1 /
C A N /C S A C 22. 2 N O.6 01. 1
9F 03
U
R
US
Made in Germany
P eissenberg
MMMMM-Y YY Y
FireWire
Connector
Direction
Figure 40
CALIBRATION:
(1)
Edition 1, Revision 3
10-2009
CR 30-X
Type 5175 / 100/110
Chapter 3.6 / 27
DD+DIS150.06E
2.5.3
Verification
(2)
(3)
(4)
(5)
Result
Edition 1, Revision 3
10-2009
CR 30-X
Type 5175 / 100/110
Chapter 3.6 / 28
DD+DIS150.06E
Calibrations
3.1
Shading Calibration
Purpose
Use Case
Determine the sensitivity of the scanning system for each pixel of the scan line:
A flatfield is scanned. For each pixel a sensitivity factor is stored in an EEPROM
on the PMT.
Perform a shading calibration in following cases:
Whenever the PMT or light collector have been removed (e.g. for replacing the
slow scan module)
Whenever stripes in slow scan position are visible, which cannot be removed by
cleaning of the scan line with the cleaning brush
REQUIRED TOOLS:
1,5 mm Copper Filter*
CM+9 5155 1015 2
* The last digit in the spare part number indicates the spare part revision at release of
this document. When ordering, the actual revision of the spare part is delivered.
REQUIRED TIME:
60 minutes
(1)
Select the largest available cassette. If the largest available cassette is only a
Full Leg / Full Spine (FLFS) Cassette, use this cassette.
In this case see the special instructions for Shading Calibration with a FLFS
cassette in section 3.1.3.
(2)
Open the cassette and check the image plate (IP) for scratches. If no scratches
are visible, select this IP.
(3)
(4)
Mount the Cu-filter below the collimation window at the X-ray device.
Edition 1, Revision 3
10-2009
CR 30-X
Type 5175 / 100/110
Chapter 3.6 / 29
DD+DIS150.06E
(5)
X-Ray tube
Rotating
Anode
517536aa.cdr
NOT OK
OK
Figure 41
(6)
Exposed
area
517536ab.cdr
Cassette
NOT OK
OK
Figure 42
(7)
12 mAs
75 kV
1.3 m (51,2) distance
Large focus
(8)
(9)
Edition 1, Revision 3
10-2009
CR 30-X
Type 5175 / 100/110
Chapter 3.6 / 30
DD+DIS150.06E
(16) Wait until the calibration data is transferred. In case of an error refer to section
3.1.2, Possible Reasons for a failed Shading Calibration.
(17) Remove cassette from digitizer.
NOTE:
The digitizer records 6 shading calibration lines (Gain 1 to Gain 6), one for
each facet of the polygon mirror.
Additionally to the storage on the PMT, the calibration file is stored in the
following folder:
C:\Documents and Settings\All Users\Application
Data\Agfa\CR30-X\Gains00.stc
The calibration file is overwritten by the new one. To keep a certain
calibration file, save it with a different name.
Edition 1, Revision 3
10-2009
CR 30-X
Type 5175 / 100/110
Chapter 3.6 / 31
DD+DIS150.06E
3.1.1
Verification
(2)
(3)
Check the image for homogeneity in slow scan and fast scan direction.
Evaluation:
If the scanned image does not show stripes in fast scan or slow scan direction or
large area inhomogeneities no further action is required.
If there are unacceptable effects in slow scan or fast scan direction, see
Troubleshooting Guide, DD+DIS151.09E.
NOTE:
Checking the images may only happen on light boxes and viewing stations that
comply with the specified installation and configuration conditions for diagnostic
reading. For details please refer to the respective instructions of the viewing stations
or light boxes.
The slow scan direction is always the long side of the scanned area.
43 cm
24 cm
18 cm
Slow scan
direction
35 cm
30 cm
24 cm
30 cm
15 cm
517509ab.cdr
Result
Edition 1, Revision 3
10-2009
CR 30-X
Type 5175 / 100/110
Chapter 3.6 / 32
DD+DIS150.06E
3.1.2
3.1.3
PMT exposed to direct sunlight before calibration (in this case PMT has lower
sensitivity)
Appendix: Changes in the Shading Calibration Procedure if Full Leg / Full Spine
Cassettes are in Use
NOTE:
This section describes the changes in the shading calibration procedure, if a FLFS
cassette is used. See section 3.1.
Reason for different proceeding: In opposite to Genrad (General Radiology)
cassettes, FLFS cassettes do not have the complete IP area covered with a
backscatter protection: The lead foil does not cover approximately 10 mm on the
upper and lower side of the cassette.
Genrad Cassette
FLFS Cassette
Backscatter
Protection
FLFS
10 mm
10 mm
FLFS
Figure 45
Edition 1, Revision 3
10-2009
CR 30-X
Type 5175 / 100/110
Chapter 3.6 / 33
DD+DIS150.06E
(1) Put 2 cassettes underneath the FLFS cassette when using a FLFS cassette for
shading calibration.
NOTE:
This covers the area where the cassette has no backscatter protection.
Only this area has to be covered, which is used by the digitizer to
determine the shading calibration curves.
FLFS cassette
Area used by
shading calibration
Area used by
shading calibration
(2) Erase the two cassettes which were put underneath the FLFS cassette during
exposure.
(3) Observe IMPORTANT note below when evaluating the flatfield
(Evaluating the flatfield is described in section 3.1.1).
IMPORTANT:
When using a FLFS cassette for flatfield
exposure, the image will be slightly
darker in the upper and lower image
area where the backscatter protection is
removed (approximately 1 cm each).
These darker zones also have to be
used for image quality evaluation in
slow scan and fast scan direction to
verify a successful shading calibration.
Flatfield with
FLFS Cassette
Flatfield with
Genrad Cassette
Effect exaggerated
Figure 48
Edition 1, Revision 3
10-2009
CR 30-X
Type 5175 / 100/110
Chapter 3.6 / 34
DD+DIS150.06E
3.2
Stall Calibration
Purpose
The stall calibration determines the optimum motor current for stall detection:
The motor drives several times to its home and end position with different motor
currents. The mean value of these different motor currents is used as motor control
current, where the steps of the motor are in the specified, allowed range.
It compensates the differences (e.g. friction; motor controller) of an assembly in series
production.
Use Case
After replacement of the IP handling board (as this has the motor controllers
built in)
REQUIRED TOOLS:
CR 30-X service client on processing station
REQUIRED TIME:
Approximately 1 minute per stall calibration
Result
(1)
(2)
(3)
(4)
(5)
(6)
Edition 1, Revision 3
10-2009
CR 30-X
Type 5175 / 100/110
Chapter 3.6 / 35
DD+DIS150.06E
NOTE:
The stall calibration does not harm - it can be performed anytime. It is required
however only in the cases as described previous page.
Edition 1, Revision 3
10-2009
CR 30-X
Type 5175 / 100/110
Chapter 3.6 / 36
DD+DIS150.06E
Chapter 3.7
The digitizer CR30-X does not have device specific settings which may be
altered by the service engineer.
All settings are either determined in production and stored in E-Lables in the
corresponding module, or - at replacement of parts - have to be determined
again by an adjustment routine. These new values are automatically written to
the corresponding E-Label.
Edition 1, Revision 0
CR30-X
Type 5175/100
Chapter 3.7 / I
Agfa Company Confidential
Chapter 3.8
HEALTHCARE
Imaging Services
CR 30-X
Type 5175 / 100/110
Document History
Edition.
Revision
Release
Date
Changes
compared to previous Version 1.1:
1.2
04-2008
Referenced Documents
Document
Title
DD+DIS063.08E
Edition 1, Revision 2
04-2008 printed in Germany
DD+DIS150.06E
Manufacturer
Agfa HealthCare N.V.
Publisher
Agfa-Gevaert HealthCare GmbH
Tegernseer Landstrae 161
D - 81539 Mnchen
Germany
Copyright 2008 Agfa HealthCare N.V.
All rights reserved.
Technical modifications reserved.
AGFA and the Agfa-Rhombus are trademarks of Agfa HealthCare N.V.
WARNING:
Improper operation or service activities may cause damage or injuries.
INSTRUCTION:
Strictly observe all safety directions within the "Generic Safety Directions" and on
the product.
NOTE:
To verify the latest version of single documents and of Service Manuals refer to the
Document Type Order List in the GSO library.
Edition 1, Revision 2
04-2008
CR 30-X
Type 5175 / 100/110
Chapter 3.8 / 2
Agfa Company Confidential
DD+DIS150.06E
LIST OF CONTENTS
Supported Functions.................................................................................................................4
1.2
Limitations .................................................................................................................................4
7.2
8.1.1
8.1.2
8.2
8.2.1
8.2.2
8.2.3
After software Upgrade Image Plate transport stops before Scan Unit due to corrupted Slow
Scan Board E-Label (HQ_0708220004).................................................................................15
8.2.4
8.3
Edition 1, Revision 2
04-2008
CR 30-X
Type 5175 / 100/110
Chapter 3.8 / 3
Agfa Company Confidential
DD+DIS150.06E
1.1
Release Date
09/2006
Digitizer SNs
1001
Availability
Prerequisites
None
Supported Functions
All functions as specified in the Design Input Requirements i.e.
Open the inserted cassette (15 x 30 cm; 18 x 24 cm; 24 x 30 cm; 35 x 43 cm) and
transport the IP to the Scan Unit.
Acquire the ID data via fast ID mode from the processing station.
Scan (= stimulate and scan) the image on the IP.
Send the raw image to the processing station
Erase the IP.
Put the IP back in its cassette, close cassette
Update ID data on IP (cycle counter).
1.2
Limitations
Shading calibration curves are stored at the processing station only. This means, if
a digitizer is operated with another processing station without restore, its shading
calibration is lost
Edition 1, Revision 2
04-2008
CR 30-X
Type 5175 / 100/110
Chapter 3.8 / 4
Agfa Company Confidential
DD+DIS150.06E
Software Upgrade
Release Date
11/2006
Digitizer SNs
1021
Availability
Prerequisites
None
*Note:
2.1
Edition 1, Revision 2
04-2008
CR 30-X
Type 5175 / 100/110
Chapter 3.8 / 5
Agfa Company Confidential
DD+DIS150.06E
Software Upgrade
Release Date
01/2007
Digitizer SNs
1021
Availability
Prerequisites
*Note:
3.1
Edition 1, Revision 2
04-2008
CR 30-X
Type 5175 / 100/110
Chapter 3.8 / 6
Agfa Company Confidential
DD+DIS150.06E
Software Update
Release Date
02/2007
Digitizer SNs
Availability
Prerequisites
*Note:
4.1
Edition 1, Revision 2
04-2008
CR 30-X
Type 5175 / 100/110
Chapter 3.8 / 7
Agfa Company Confidential
DD+DIS150.06E
Software Upgrade
Release Date
05/2007
Digitizer SNs
Availability
Prerequisites
None
*Note:
5.1
Edition 1, Revision 2
04-2008
CR 30-X
Type 5175 / 100/110
Chapter 3.8 / 8
Agfa Company Confidential
DD+DIS150.06E
Software Update
Release Date
08/2007
Digitizer SNs
2292 (+ 2268)
Availability
Prerequisites
*Note:
6.1
Solved sporadic error codes 10754, 10761 or 10767 may be displayed during
boot-up or just before scanning. This fault mainly happened with machines that
have been upgraded from software CRE_1206X to CRE_1309 in the field.
Edition 1, Revision 2
04-2008
CR 30-X
Type 5175 / 100/110
Chapter 3.8 / 9
Agfa Company Confidential
DD+DIS150.06E
Software Upgrade
Release Date
11/2007
Digitizer SNs
Availability
Prerequisites
*Note:
7.1
Edition 1, Revision 2
04-2008
CR 30-X
Type 5175 / 100/110
Chapter 3.8 / 10
Agfa Company Confidential
DD+DIS150.06E
7.2
Edition 1, Revision 2
04-2008
CR 30-X
Type 5175 / 100/110
Chapter 3.8 / 11
Agfa Company Confidential
DD+DIS150.06E
Software Upgrade
Release Date
04/2008
Digitizer SNs
Availability
Prerequisites
*Note:
8.1
8.1.1
8.1.2
Software CRE_1509 supports both, the existing PMI board CM+9 5175 2290 X*
and the new developed PMI2 board CM+9 5175 2180 X (which is not compatible
to software < CRE_1509).
* The 'X' is a placeholder for the digits '0-9' which indicate the latest version obtainable in the spare
part supply system. When ordering a spare part please replace the 'X' by a '0'. You will always
receive the latest obtainable version.
Software CRE_1509 displays error 4642 after replacement of a spare part with
E-label (i.e. each spare part which keeps spare part specific data):
This error can only be solved in service client menu Configuration Exchanged
Edition 1, Revision 2
04-2008
CR 30-X
Type 5175 / 100/110
Chapter 3.8 / 12
Agfa Company Confidential
DD+DIS150.06E
Components. There you have to enter, whether the spare part is a complete
module, or whether it is a printed circuit board only.
Example:
- For replacement of slow scan board select Board.
- For replacement of slow scan unit (incl. slow scan board) select Module.
This new feature was required due to logistical reasons. For details refer to
Technical Documentation, chapter 3.2, Tools.
Figure 1
NOTE:
All modules with E-labels have installation instructions enclosed, which describe the
correct proceeding.
8.2
8.2.1
Symptom
Defect
Software problem
Solution
Edition 1, Revision 2
04-2008
CR 30-X
Type 5175 / 100/110
Chapter 3.8 / 13
Agfa Company Confidential
DD+DIS150.06E
8.2.2
Symptom 1
After replacement of an IP Handling Board (CM+9 5175 5230 X) it is possible that after
insertion of the cassette, during cassette clamping error 4152 pops up "Cassette fixer
detected unknown format".
Symptom 2
After replacement of a Slow Scan Board (CM+9 5175 2120 X) it is possible that the
aspect ratio of the scanned image is not exactly 1:1 anymore, but e.g. 1:0,99 or 1:1,01.
IMPORTANT:
Both symptoms only appear, if both of following conditions are fulfilled:
The digitizer at the production site which is used for board testing and the digitizer
on site, where the new boards are built in, deviate a lot in mechanical
tolerances concerning cassette clamping or the slow scan gear box.
The NX processing station is always switched on
Defect
Solution
Edition 1, Revision 2
04-2008
CR 30-X
Type 5175 / 100/110
Chapter 3.8 / 14
Agfa Company Confidential
DD+DIS150.06E
8.2.3
After software Upgrade Image Plate transport stops before Scan Unit due to
corrupted Slow Scan Board E-Label (HQ_0708220004)
Symptom
In some cases after software upgrade to CRE_1309 or exchange of the slow scan
board the digitizer stops scanning when the first image plate enters the scan unit, due
to a corrupted slow scan E-Label:
In this case the digitizer waits for information which is normally stored in the E-label.
No error message is issued. The digitizer boots-up properly, as the corrupted slow scan
board e-label parameters are not used during boot-up. With a reset the image plate is
returned in the cassette. Scanning the next image plate gives the same result.
Defect
Solution
8.2.4
In some cases, when working with the Service Client, the error viewer was not
visible, if an error occurred (e.g. in diagnose cycle). To solve the problem in the
worst case a reset of the control PC was required. In SW 1509 always the error
viewer is visible if an error occurs while working with the service client.
Total number of cycles of the product was not written to the E-label and the
infocounter.
Stall Calibration details were not logged in Service Activity Log file
Edition 1, Revision 2
04-2008
CR 30-X
Type 5175 / 100/110
Chapter 3.8 / 15
Agfa Company Confidential
DD+DIS150.06E
In some cases the service client could not re-establish communication after
software upload. It was required to reboot the control PC.
If the digitizer had problems writing to the cassette chip, not only NX message
119: The file could not be saved appeared, but also a false error message
Problem during scan: The image may not have the required quality.
During boot-up the stepper motors did not retry to go to home position if a problem
occurred. Now three retries are performed.
Retry sequence of cassette opener did not work, if cassette cover was
mechanically blocked.
Some timeouts during scan were not transferred to the control PC, e.g. if scan
command did not return a response.
8.3
In software CRE_1405 it was possible that regular, bright, wide stripes in fast scan
direction appeared, if first a 35x43 cm Genrad scan followed by a 35x43 cm FLFS
(Full leg / full spine) scan a was performed.
Open Issues
In software CRE_1509 it is erroneously possible to perform the service function
Diagnose Cycle and Endurance Run remotely. In previous software versions these
two functions were blocked for service connections via remote desktop.
CAUTION:
A local service engineer may work at the device while a remote service engineer
executes a Diagnose Cycle: Risk of injury.
Do not execute service function Diagnose Cycle via Remote Desktop connection.
Edition 1, Revision 2
04-2008
CR 30-X
Type 5175 / 100/110
Chapter 3.8 / 16
Agfa Company Confidential
Chapter 3.9
HEALTHCARE
Imaging Services
CR 30-X
Type 5175 / 100/110
Please note
The Chapter FAQ, Edition 1, Revision 0, has been replaced by the separately
available Troubleshooting Guide, DD+DIS151.08E
Please download the Troubleshooting Guide from the GSO Library:
MEDNET GSO Computed Radiography CR Digitizers
CR 30-X Trouble shooting guide
Edition 1, Revision 1
07/2008 printed in Germany
Agfa Company Confidential
DD+DIS150.06E
Edition 1, Revision 1
07/2008
CR 30-X
Type 5175 / 100/110
Chapter 3.9 / 2
Agfa Company Confidential
DD+DIS150.06E
List of Contents
1.1
1.2
Motors ......................................................................................................... 1
1.3
1.4
Connectors ................................................................................................. 2
Edition 1, Revision 0
CR30-X
Type 5175/100
Chapter 4 / I
Agfa Company Confidential
DD+DIS150.06E
Reference Diagrams
1.1
1.2
Refer to
Figure 3, page 5
Figure 3, page 5
Figure 3, page 5
Motors
Name
Refer to
Figure 2, page 4
Figure 2, page 4
Figure 2, page 4
Figure 2, page 4
Figure 2, page 4
Figure 2, page 4
Figure 2, page 4
Figure 2, page 4
Edition 1, Revision 0
CR30-X
Type 5175/100
Chapter 4 / 1
Agfa Company Confidential
DD+DIS150.06E
1.3
1.4
Refer to
Power Supply
Figure 3, page 5
IP Handling Board
Figure 1, page 3
Chip Reader
Figure 3, page 5
User Interface
Figure 1, page 3
Figure 1, page 3
PMI Board
Figure 1, page 3
Roller IP Sensor
Figure 3, page 5
Figure 3, page 5
Distributor Board
Figure 3, page 5
Calibration Board
Figure 3, page 5
Figure 1, page 3
Erasure Unit
Figure 3, page 5
Roller IP Sensor
Figure 3, page 5
Connectors
Name
Refer to
Interconnector XK3
Figure 3, page 5
Chapter 4 / 2
Agfa Company Confidential
CR30-X
Type 5175/100
Edition 1, Revison 0
DD+DIS150.06E
PMI Board
PMI Power
Disribution Board
Slow-Scan Board
User Interface
517504aa.cdr
IP Handling Board
Figure 1
Edition 1, Revision 0
CR30-X
Type 5175/100
Chapter 4 / 3
Agfa Company Confidential
DD+DIS150.06E
Opener
Stepper Motor
(M3)
Drawer
Stepper Motor
(M4)
Fixer
Stepper Motor
(M1)
Slow Scan
Stepper Motor
(M9)
Locking
Stepper Motor
(M17)
Roller
Stepper Motor
(M5)
517504ab.cdr
Roller Clamping
Stepper Motor
(M6)
Figure 2
DOCUMENT CONTROL NOTE:
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Chapter 4 / 4
Agfa Company Confidential
CR30-X
Type 5175/100
Edition 1, Revison 0
DD+DIS150.06E
Calibration Board
Distributor Board
Chip Reader
Firewire Cable
Interconnector
XK3
Power Supply
Roller IP Sensor
514504ac.cdr
Figure 3
DOCUMENT CONTROL NOTE:
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 0
CR30-X
Type 5175/100
Chapter 4 / 5
Agfa Company Confidential
DD+DIS150.06E
Frame
Sheet 1 of 9
Cassette Unit
Sheet 2 of 9
Scan Modul
Sheet 3 of 9
Light Collector
Sheet 4 of 9
Scanner Unit
Sheet 5 of 9
Erasure Unit
Sheet 6 of 9
Sheet 7 of 9
Sheet 8 of 9
Sheet 9 of 9
Edition 1, Revision 0
CR30-X
Type 5175/100
Chapter 4 / 7
Agfa Company Confidential
+24VDC_ PMI_POWER_DISTR:
-W924,-W925,-W926,-W927
RD,RD,RD,RD
#13127
/Light_Collector.1A
+24VDC_CASSETTE_UNIT
-W932,-W933,-W934,-W935
RD,RD,RD,RD
#13126
/Cassette_Unit.3A
+24VDC_SCANNER_UNIT
-W940,-W941,-W942,-W943
RD,RD,RD,RD
#13128
/Scanner_Unit.2A
-W948,-W949,-W950,-W951
RD,RD,RD,RD
110-240VAC 50/60Hz
Main-Switch
-S3
-BU139
F8.5175.8585.0A
1/1
1/1
-BU140
L1
N1
-BU135
-3
1/1
1/1
F7.0475.9990.0
1/1
-W968
BU
-W972
BK
-4
-BU136
-BU138
1/1
-BU141
1/1
1/1
F7.0475.9990.0
-5
-W970
GNYE
-W971
BU
ACL
+24VDC
ACN
Power Supply
24V DC / 12,5A +24VDC
PE
1/1
F7.0475.9990.0
F7.0486.1420.0
-W972
BK
GND
GND
PE1 1/1
-2
-W963
RD
-W964
RD
1/1
F7.0473.6252.0
-6
1/1
F7.0473.6252.0
1/1
-S2
-BU70
1/1
RD,RD
1/1
-W962
RD
-BU71
1/1
F7.0473.6252.0
1/1
1a
1d
1b
1e
1/1
1c
1f
1/1
2a
2d
2b
2e
1/1
2c
2f
1/1
3a
3d
3b
3e
1/1
3c
3f
1/1
4a
4d
4b
4e
4c
4f
-BU91
1/1
-W965
BK
-7
1/1
-BU98
-BU104
1/1
-W944,-W945
BK,BK
F8.5175.8580.0
-W966
BK
1/1
-BU111
-W930,-W931
BK,BK
-PE
1/1
F7.0471.3217.0
1/1
-W932,-W933
RD,RD
-W934,-W935
RD,RD
1/1
1/1
1/1
-W948,-W949
RD,RD
-W950,-W951
RD,RD
-BU101
1/1
1/1
-BU107
1/1
-BU108
-W952,-W953
BK,BK
-W954,-W955
BK,BK
1/1
-W936,-W937
BK,BK
1/1
-W938,-W939
BK,BK
-BU109
-BU110
/ERASURE_UNIT.2A
-BU95
-BU100
-BU103
-W928,-W929
BK,BK
1/1
-BU106
-BU105
1/1
F7.0473.6252.0
1/1
-BU94
-BU102
F8.5175.8580.0
-BU93
-BU99
-BU97
-W942,-W943
RD,RD
F8.5175.8580.0
-1
-BU92
-BU96
-W940,-W941
-BU73
#13125
-XK3
F7.0473.6056.0
-BU90
-W961
RD
-BU69 7
1/1
-B420
C101837
F7.0460.9064.0
-W926,-W927
RD,RD
F8.5175.8580.0
-W967
BK
-BU137
PE
-W924,-W925
RD,RD
-BU112
1/1
-W946,-W947
BK,BK
-W946,-W947
BK,BK
#13125
#13126
Input Firewire
-W928,-W929,-W930,-W931
BK,BK,BK,BK
#13127
-W944,-W945,-W946,-W947
BK,BK,BK,BK
#13128
GND_ERASURE_UNIT
GND_CASSETTE_UNIT
/Cassette_Unit.3A
GND_PMI_POWER_DISTR:
/Light_Collector.1A
GND_SCANNER_UNIT
/ERASURE_UNIT.2A
/Scanner_Unit.2A
F7.0477.1087.0 6p FireWireCable
#14073
/Light_Collector.3A
4
4
-ST33
F7.0471.9197.0
-W960
F7.0477.1087.0
USER
APPR.
NAME
03.01.2006
Leith.
General Schematic
CR30-X
Sheet name:
FRAME
39319
29.08.06
Leith.
0A
38631
31.05.06
Leith.
37407
20.04.06
Leith.
F1.5175.8503.0
Change-NR
DATE
NAME
Replacement for
IDX
NORM
DATE
0B
AGFA
+24VDC_ERASURE_UNIT
Project
5175
/ 100
Sheet 1
of
9 SH
USER_INTERFACE
F8.5175.1580.X
Erasure
BN
1/2
BK
2/2
L7.0471.9060.0
/FRAME.1E
/FRAME.4F
-BU79
*8/8
L7.0471.9093.0
#13126
*10/10
F7.0429.0008.0
LockingHomePos
-BU170
-BU80
T1A
T1A
+24VDC
User Interface
-S1
+5V
+5V_RF
-P27
*8/8
F7.0429.0121.0
IP-Handling-Board
F8.5175.5230.X
-P4
T0,5A
T2,5A
-P28
*10/10
F7.0429.0008.0
*4/4
L7.0429.0009.0
-P10
*4/4
L7.0429.0009.0
-P9
*4/4
L7.0429.0009.0
-P8
*4/4
L7.0429.0009.0
-P5
*4/4
L7.0429.0009.0
-P3
*4/4
L7.0429.0009.0
-P14
*4/4
L7.0429.0009.0
-P13
2/2
F7.0429.0044.0
*8/8
F7.0429.0040.0
*10/10
LockingCassetteContact
CoverOpenerControl
DrawerHomePos
OpenerHomePos
LockingEndPos
FixerHomePos
AlignmentHomePos
*4/4
L7.0429.0009.0
+39VDC
CassetteEdgeSensor
F8.5175.8510.0
*10/10
F7.0471.8178.0
-P1
-P12
LD1
-W922
F8.5175.8640.0
*4/4
L7.0429.0009.0
-P15
1/2
Reset
-F1#2
1/1
F8.5175.8640.0
1/1
F8.5175.8630.0B
POWER
-P16
-W919
-F1#1
-BU4
-BU78
Supply on BU170
toleranc limit +/- 5%
1,2 = +24VDC
8,9, = GND
-F10
-W921
-BU5
-ST5
-BU77
1/6
2/6
3/6
4/6
-S1
L7.0426.0950.0
*4/4
F7.0429.0036.0
*4/4
F7.0471.8339.0
-W918
-ST11
E-Label
1/6
2/6
3/6
4/6
5/6
6/6
F7.0429.0033.0
-P18
-BU1#14
2
Reserve
-BU76
F8.5175.8620.0
1/6
2/6
3/6
4/6
-W917
Supply on St10
toleranc limit +10% / -5%
4 = +24VDC
3 = GND
-B2
f7-9499-6941-1.dfb
-ST10
F8.5175.8610.0
-BU1
+F7.0486.1401.0
*4/4
1/6
2/6
3/6
4/6
5/6
6/6
F7.0429.0033.0
*4/4
F7.0471.8344.0
F8.9499.6940.X
gn
rd
RollerIpSensor
-F8
CHIP READER
LED
-F7
24VDC
/ERASURE_UNIT.2A
DIL-SWITCH
4
OFF
NC
T3,5A
F8.5175.8590.0
*8/8
F7.0471.8341.0
-S2
Switch -S2
ON = start application
Off = start debug monitor
-W916
-BU87
NC
Sheet 1
3
Supply on Bu79
toleranc limit +/1,2,3,4 = +24VDC
5,6,7,8 = GND
-ST1#3
*10/10
F7.0429.0023.0
CAN-Terminating
resistor 121 Ohm
Bootload / Debug
-P17
-P32
BK
-M1
F7.0430.8362.0
SM-Fixer
-M17
F7.0430.8364.0
SM-Unlocking
-M3
F7.0430.8357.0
SM-Opener
SM-Drawer
BK
-M6
F7.0430.8357.0
-M5
F7.0430.8357.0
-M4
F7.0430.8364.0
YE
YE
BK
WHRD
WHYE
RD
BK
WHGN
WHBK
1/4
2/4
3/4
4/4
F7.0429.0039.0
1/8
2/8
3/8
4/8
5/8
6/8
7/8
8/8
F7.0429.0040.0
-P29
-BU86
F8.5175.8680.0
YE
YE
RD
WHRD
WHYE
BK
BU
WHGN
WHBK
WHGN
WHBK
YE
WH
RD
1/8
2/8
3/8
4/8
5/8
6/8
7/8
8/8
L7.0471.9093.0
1/8
2/8
3/8
4/8
5/8
6/8
7/8
8/8
F7.0429.0040.0
-BU85
F8.5160.4519.0
YE
F8.5175.8670.0
YE
YE
RD
WHGN
WHBK
BU
WHRD
WHYE
BK
GN
YE
WH
RD
1/8
2/8
3/8
4/8
5/8
6/8
7/8
8/8
L7.0471.9093.0
-P25
1/4
2/4
3/4
4/4
F7.0429.0039.0
1/4
2/4
3/4
4/4
L7.0471.9119.0
-P31
-BU84
1/8
2/8
3/8
4/8
5/8
6/8
7/8
8/8
F7.0429.0040.0
-P19
1/8
2/8
3/8
4/8
5/8
6/8
7/8
8/8
L7.0471.9093.0
-BU83
F8.5160.4519.0
YE (BU)
GN
-B419
F7-5175-5231-A.dfb
Reserve
GN
BK
SM-Roller-Clamping
SM-Roller
USER
APPR.
AGFA
SM-Alignment
GN
WHRD
WHYE
RD
YE
F8.5175.8670.0
F8.5175.8660.0A
BK
-BU81
CAN BUS
1/4
2/4
3/4
4/4
F7.0429.0039.0
#9/11
F7.0429.0129.0
-P30
-P17
1/4
2/4
3/4
4/4
L7.0471.9119.0
#9/10
F7.0471.8924.0
-BU82
-BU131
YE
/Light_Collector.4A
F8.5175.8540.0
1/8
2/8
3/8
4/8
5/8
6/8
7/8
8/8
L7.0471.9093.0
-P23
-W924_2
1/8
2/8
3/8
4/8
5/8
6/8
7/8
8/8
F7.0429.0040.0
#2/11
F7.0429.0129.0
NAME
03.01.2006
Leith.
General Schematic
CR30-X
Sheet name:
Cassette_Unit
0B
39319
29.08.06
Leith.
0A
38631
31.05.06
leith.
37407
20.04.06
Leith.
F1.5175.8503.0
Change-NR
DATE
NAME
Replacement for
IDX
NORM
DATE
Project
5175
/ 100
Sheet 2
of
9 SH
/Light_Collector.1F
Mirror
NC
NC
*2/2
F7.0471.1749.0
*10/10
GENERIC10
*8/8
GENERIC8
-BU169
*8/8
F7.0471.9053.0
-ST1
12, 25 = GND
13 = +12V
24 = -12V
S/26
-ST1
-BU130
#25/26 #24
F7.0471.8879.0
-ST5
*10/10
F7.0429.0008.0
#25/26
F7.0429.0134.0
-W1065
F8.5175.2980.0
#6
LINCOLN
*10/10
F7.0471.8820.0
-BU171
*4/4
F7.0429.0036.0
-B317
F7-5170-2931-C.dfb
BL317
-W1066
F8.5175.2970.0
#10
-W1067
COPAL
F8.5175.2960.0
#7
USER
APPR.
NAME
03.01.2006
Leith.
General Schematic
CR30-X
Sheet name:
Scan_Modul
39319
29.08.06
Leith.
0A
38631
31.05.06
Leith.
37407
20.04.06
Leith.
F1.5175.8503.0
Change-NR
DATE
NAME
Replacement for
IDX
NORM
DATE
0B
AGFA
-B319
Polygon PCB
BL319
-CN1
E-Label
LINCOLN
*2/2
L7.0471.8947.0
-J5
+10V MP3
-BU1
-B320
F1-5170-2913-A.dfs
BL321
-BU147
-J1
IC3
MP2
-BU148
IC1
1, 14 = GND (+24V)
2 = +24V
*10/10
F7.0471.8178.0
Polygon
-5V
Supply on BU130
tolerance limit +-5%
-BU172
*12/12
F7.0429.0032.0
1, 2, 13 = GND
3,4 = +5V (+-2%)
5, 6 = -5V (*-2%)
12 = +10V (+-1%)
MP1
*12/12
F7.0429.0032.0
-ST3
IC2
*12/12
F7.0471.8556.0
LD - MODULATOR - CR20
F8.5175.2910.X
Optical lense
Supply on BU50
-ST7
+5V
LD current control
-W1071
*16/16
F7.0429.0029.0
LD not connected
LED1
F7.5175.8515.0
*8/8
F7.0429.0121.0
-ST2
-BU149
*16/16
F7.0471.8395.0
*8/8
F7.0429.0121.0
-BU51
F8.5175.2920.0
-ST6
Optical lense
*16/16
F7.0471.8395.0
*16/16
F7.0429.0029.0
-W304
-BU50
*8/8
F7.0471.8341.0
-ST1
D12
-ST3
-ST4
*8/8
F7.0429.0121.0
Laserdiode
Project
5175
/ 100
Sheet 3
of
9 SH
-L2#12
T2,5A
T2,5A
T1A
-F20
-F1
1,5VD
*50/50
L7.0472.1012.0
-BU178
*50/50
L7.0472.1012.0
#50
-P36
*50/50
F7.0471.9202.0
+20V PMT
+20V PMT
+8V Disp.
-12V OPT
Power Supply
Polygon Motor
FPGA NRESET
-LD30
5.0VD
-LD45
M32 PMT + OK
-LD46
M32 PMT - OK
-LD47
M32 Analog OK
-LD48
*80/80
-S2
OFF
DIL-SWITCH
-S4
OFF
DIL-SWITCH
OFF
-S3
-LD36
M32 Reserve 2
FPGA Test 4 -LD34
M32 CAN OK
M32 NCS 2
-LD39
-LD50
M32 LCD OK
M32 NCS 3
-LD40
-ST1#1
*10/10
-LD2
M32 WDG
-P1
-LD31
3,3VD
-LD51
ConF/FPGA done
-F1#5
VCC_NODE
-P21
*9/9
-P30
RS232 Debug-Interface
-P34
*10/10
EMI-BOX for
PMI-BOARD
2
-P16
PMI-BOARD
#9/10
F7.0471.8670.0
CAN-Terminating
resistor 121 Ohm
-B410
4991.DFB
+24.0V POL
5.0VD_FW
-R1229
-R1#12
F8.5175.8550.0
X
/Scanner_Unit.1F
F8.5175.2290.X
-W906
-ST17
*9/9
*10/10
125mA
Test connector
*10/10
-LD38
*80/80
/Scan_Modul.1F
E-Label connector
-P1#5
M32 NCS 1
-P22
-LD33
FPGA Test 3
-LD37
*4/4
1,25A
NC
300mA
P1
3
+B422
PMT
-W960:B
-BU1
F8.5175.3040.X
-ST34
9pol. DSUB
#14073
/FRAME.4F
-B422
-W924_1
/Cassette_Unit.3A
/Scanner_Unit.2A
*4/4
M32 NCS 0
-ST23
-B411
F7-5170-4981-0.dfb
-BU177
3,3VD
-F24
-F27
-P5
*50/50
F7.0471.9202.0
+8V Analog
T2A
PWR_CAN
-LD52
F7.5175.8515.0
#25/26
-BU1#10
*17/17
F7.0471.9196.0
-F19
F8.5175.8535.0A
#25/26
-W1071
-BU176
-P25
T1A
-F18
-W1070
S/26
-P17
+B422
T1A
-F16
+12V OPT
+12V
*6/6
T3,5A
-F15
T1A
-P2#11
T3,5A
-F23
*6
P323326
Supply on Bu129
tolerance limit + / - 5%
1,2,3,4= +24VDC
5,6,7,8 = GND
-F1#1
T2,5A
V33_PHY
DIL-SWITCH
V33_PHY#12
-S1
-L1#12
24VDC
*8/8
F7.0429.0038.0
S/11
S/11
T5A
#9/11
-F13
1,5VD
-P1#6
-F25
F8.5170.4980.X
/FRAME.1E
/FRAME.4F
*8/8
#13127
F7A
-P3
-LD32
#9/10
F7.0471.8924.0
IN 24VDC
-BU129
-BU133
T1A
F8.5175.8510.0
-P17
PMI_POWERDISTRIBUTION-BOARD
Reset
F8.5175.8540.0
Light collector
4
USER
APPR.
NAME
03.01.2006
Leith.
General Schematic
CR30-X
Sheet name:
Light_Collector
39319
29.08.06
Leith.
0A
38631
31.05.06
Leith.
37407
20.04.06
Leith.
F1.5175.8503.0
Change-NR
DATE
NAME
Replacement for
IDX
NORM
DATE
0B
AGFA
Auto_Cal_LED
Project
5175
/ 100
Sheet 4
of
9 SH
SlowScanliftPosSensor
*4/4
-BU146
#4
-W1063
*4/4
-BU63
-B413
f7-9499-5881-0.dfb
-ST10
F8.9499.5880.X
-W1062
-P2
*4/4
F7.0471.8339.0
*4/4
-W906
-BU144
F8.5175.8550.0
*10/10
F7.0471.8178.0
-P3
*4/4
L7.0429.0009.0
*10/10
F7.0429.0008.0
/Light_Collector.2F
to PMI_P16
-W1064
-BU143
F8.5175.2415.0
*10/10
F7.0471.8178.0
*10/10
F7.0429.0023.0
-P7
-ST1#3
*8/8
F7.0429.0040.0
*4/4
F7.0429.0036.0
-BU1#4
-P3
-F1
-P4
*10/10
F7.0429.0008.0
-B421
F7-5175-1561-A.dfb
BL331
*4/4
L7.0429.0009.0
-BU64
#4
T1A
*4/4
L7.0429.0009.0
-BU145
F8.5175.8560.0
F8.5175.8510.0
-P1
*4/4
F7.0471.8339.0
F8.5175.2410.0
Supply on Bu67
tolerance limit + / - 5%
1,2,3,4= +24VDC
5,6,7,8 = GND
Distributor Board
F8.5175.1560.X
24VDC
Reset
SLOW-SCAN Board
F8.5175.2120.X
-F1#1
+5VDC
-W924_1,-W924_2
ON = start application
off = start debug monitor
NC
SM-SlowScan
A
-ST10
BOL
-BU127
*10/10
F7.0471.8178.0
-B412
F8.5175.2990.X
EOL F7-5175-2991-A.dfb
1b
2b
3b
4b
1a
2a
3a
4a
OG
YE
-BU126
RD
YE
WHGN GN
BN
WHRD
RD
-M9
F7.0430.8366.0
AGFA
F8.5175.2640.0
S7
-ST15
F7.0471.8480.0
D18 D20
1/4
2/4
3/4
4/4
F8.5175.2510.0
-BU121
-B415
F7-5175-2121-A.dfb
Calibration Board
D1
*10/10
F7.0429.0008.0
1/4
2/4
3/4
4/4
F7.0429.0039.0
#2/11
F7.0429.0129.0
D19
F8.5175.8510.0
-P5
NC
4
-S2
#9/11
F7.0429.0129.0
Power ON
1/4
2/4
3/4
4/4
F7.0429.0039.0
CAN_BUS
-P5
1/4
2/4
3/4
4/4
L7.0471.9119.0
/Light_Collector.4A
#9/10
F7.0471.8924.0
-P1
-BU132
X
-LD2
F7.0449.4004.0
F8.5175.8540.0
2
-S1
-P2
T1A
1/4
2/4
3/4
YE
4/4
L7.0471.9119.0
*8/8
F7.0429.0040.0
-BU134
-P4
1/4
2/4
3/4
4/4
*8/8
L7.0471.9093.0
OFF
#13128
DIL-SWITCH
-BU67
/FRAME.1E
/FRAME.4F
USER
-M8
F7.0430.8291.0
APPR.
SM-SlowScanLift
C
NAME
03.01.2006
Leith.
General Schematic
CR30-X
Sheet name:
Scanner_Unit
0B
39319
29.08.06
Leith.
0A
38631
31.05.06
Leith.
37407
20.04.06
Leith.
F1.5175.8503.0
Change-NR
DATE
NAME
Replacement for
IDX
NORM
DATE
Project
5175
/ 100
Sheet 5
of
9 SH
Erasure Unit
F8.5175.3990.X
-BU61
F8.5175.8510.0
/FRAME.1E
/FRAME.3F
*8/8
L7.0471.9093.0
#13125
+B414
red
LD16
LD17
red
LD31
LD32
red
LD47
LD48
red
LD62
LD63
royalblue
LD78
-P3
*8/8
F7.0429.0038.0
Supply on BU61
toleranc limit + / - 5%
1,2,3,4 = 24VDC
5,6,7,8 = GND
LD1
T1A
-F1#1
+5V
F7.0459.0304.0
T5A
-F2
24VDC
F7.0459.0314.0
-W916
/Cassette_Unit.2H
F8.5175.8590.0
-BU62
*8/8
F7.0471.8341.0
+B414
-P2
*8/8
F7.0429.0121.0
F7.0429.0044.0
2/2
LD93
2/2
L7.0471.9060.0
1/2
-P4
royalblue
BK
BK
F8.5175.3520.0
-BU175
1/2
-B414
F7-5175-3991-D.dfb
+B414
LD79
BK
S12
F7.8149.6802.0
-S4
USER
APPR.
NAME
04.01.2006
Leith.
General Schematic
CR30-X
Sheet name:
ERASURE_UNIT
39319
29.08.06
Leith.
0A
38631
31.05.06
Leith.
37407
20.04.06
Leith.
F1.5175.8503.0
Change-NR
DATE
NAME
Replacement for
IDX
NORM
DATE
0B
AGFA
BK
Project
5175
/ 100
Sheet 6
of
9 SH
SCANNER_UNIT
-BU96#1
-W940,-W941
RD,RD
-W940
RD
-BU98#1
-W942
RD
-W944
BK
-BU105#1
-W946
BK
-BU112#1
-W941
RD
-W943
RD
-W945
BK
-W947
BK
-BU67#1
1
2
3
4
5
6
7
8
BGR F8.5175.8510.0
PMI_POWER_DISTR
-BU92#1
1
-W924
RD
-BU91#1
1
-W926
RD
-W928
BK
-BU110#1
1
-W930
BK
-BU111#1
1
-W925
RD
-W927
RD
-W929
BK
-W931
BK
-BU129#1
1
2
3
4
5
6
7
8
SCANNER_UNIT
2
-BU134#1
1
2
3
4
ERASURE_UNIT
-BU99#1
1
-W948
RD
-BU100#1
1
-W950
RD
-W952
BK
-BU106#1
1
-W954
BK
-BU107#1
1
-W949
RD
-W951
RD
-W953
BK
-W955
BK
-BU123#1
1
2
3
4
-W956
BN
-W957
OG
-W958
RD
-W959
YE
-BU61#1
1
2
3
4
5
6
7
8
3
CASSETTE_UNIT
-BU93#1
1
-W932
RD
-BU94#1
1
-W934
RD
-W936
BK
-W938
BK
-BU109#1
1
-W935
RD
-W937
BK
-W939
BK
-BU79#1
1
2
3
4
5
6
7
8
4
USER
APPR.
NAME
07.02.2006
Leith.
Wire_harness_5175.8510
39319
29.08.06
Leith.
0A
38631
31.05.06
Leith.
37407
20.04.06
Leith.
F1.5175.8503.0
Change-NR
DATE
NAME
Replacement for
IDX
NORM
DATE
0B
AGFA
-BU108#1
1
-W933
RD
Project
5175
/ 100
Sheet 7
of
9 SH
BGR F8.5175.8580.0
1
-2#1
1
-W963
RD
-6#1
1
-W964
RD
-1#1
1
-W965
BK
-W965
BK
-7#1
1
-W966
BK
-W966
BK
1
-BU69#1
1
-BU70#1
-BU73#1
1
-W961
RD
-BU71#1
1
-W962
RD
1
-BU90#1
1
-BU97#1
1
-BU104#1
2
1
-BU103#1
USER
APPR.
NAME
15.02.2006
Leith.
Wire_harness_5175.8580
39319
29.08.06
Leith.
0A
38631
31.05.06
Leith.
37407
20.04.06
Leith.
F1.5175.8503.0
Change-NR
DATE
NAME
Replacement for
IDX
NORM
DATE
0B
AGFA
Project
5175
/ 100
Sheet 8
of
9 SH
BGR F8.5175.8585.0
1
-BU135#1
1
-W967
BK
-BU136#1
1
-W968
BU
-BU139#1
1
1
-3#1
1
-4#1
-W971
BU
1
-BU138#1
-BU140#1
1
-BU137#1
1
-W972
BK
-W969
GNYE
1
-BU141#1
-W970
GNYE
1
-5#1
USER
APPR.
NAME
15.02.2006
Leith.
Wire_harness_5175.8585
39319
29.08.06
Leith.
0A
38631
31.05.06
Leith.
37407
20.04.06
Leith.
F1.5175.8503.0
Change-NR
DATE
NAME
Replacement for
IDX
NORM
DATE
0B
AGFA
-PE#1
1
Project
5175
/ 100
Sheet 9
of
9 SH
CR 30-X
Type 5175/0100/0110
Internal update: 5
CONFIDENTIALITY NOTE:
Use, dissemination, distribution or reproduction of this document by unauthorized personnel is not permitted and may be unlawful.
DOCUMENT CONTROL NOTE:
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 3
01-2009 printed in Germany
DD+DIS325.06M
Manufacturer
Agfa HealthCare N.V.
Publisher
Agfa-Gevaert HealthCare GmbH
Tegernseer Landstrae 161
D - 81539 Mnchen
Germany
Copyright 2008 Agfa HealthCare N.V.
All rights reserved.
Technical modifications reserved.
AGFA and the Agfa-Rhombus are trademarks of Agfa HealthCare N.V.
WARNING:
Improper operation or service activities may cause damage or injuries.
INSTRUCTION:
(1) Read the "Generic Safety Directions" document
(see MEDNET GSO => General Info => Agfa HealthCare => Publications =>
Service Manual) prior to attempting any operation, repair or maintenance task on
the equipment.
(2) Strictly observe all safety directions within the "Generic Safety Directions" and on
the product.
WARNING:
Hazards may be introduced because of component failure or improper
operation.
INSTRUCTION:
NOTE:
To verify the latest version of single documents and of Service Manuals refer to the
Document Type Order List in the GSO library.
Edition 1, Revision 3
01-2009
CR 30-X
Type 5175/0100/0110
Chapter 5 / 2
DD+DIS325.06M
Document History
Edition.
Revision
1.3
1.3
Release
Date
Changes
compared
to previous
1.2
Changes
compared
to previous
versionversion
1.2
Contact
Spare Parts ordering
Europe
orderprocessing-europe.matrium@eads.com
Overseas
orderprocessing-overseas.matrium@eads.com
returns.matrium@eads.com
NOTE:
For Recycling Information please refer to:
http://intra.agfanet/cd/ep/ehs.nsf
Edition 1, Revision 3
01-2009
CR 30-X
Type 5175/0100/0110
Chapter 5 / 3
DD+DIS325.06M
Contents
VERKLEIDUNG
COVER
REVETEMENT
PAGES 06 - 07
SCANBAUGRUPPE
SCAN UNIT
UNITE SCAN
PAGES 08 - 09
SLOWSCAN
SLOWSCAN
SLOWSCAN
PAGES 10 - 11
KASSETTEN
CASSETTES
CASSETTES
PAGES 18 - 19
SCHUBLADE KASSETTENEINHEIT
DRAWER CASSETTE UNIT
TIROIR MODULE DE CASSETTES
PAGES 14 - 15
SCHUBLADE
DRAWER
TIROIR
PAGES 12 - 13
Edition 1, Revision 3
01-2009
CR 30-X
Type 5175/0100/0110
Chapter 5 / 4
DD+DIS325.06M
Open Bookmarks
(3)
(4)
(5)
(6)
Click on "CONTENTS"
Edition 1, Revision 3
01-2009
CR 30-X
Type 5175/0100/0110
Chapter 5 / 5
DD+DIS325.06M
01
03
09
07
08
06
05
02
04
5175_CHAP05_01.CDR
COVER
DOCUMENT CONTROL NOTE:
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 3
01-2009
CR 30-X
TYPE 5175/100/110
Chapter 5 / 6
DD+DIS325.06M
Pos. Nr. Teile-Nr.
Item no. Part No.
Pos. no. Nr. de Ref.
Benennung
Description
Denomination
CM+9517592000
CM+9517515400
CM+9517515041
CM+9517515031
CM+9517516302
CM+9517516500
CM+9517515800
USER INTERFACE
USER INTERFACE BOARD
USER INTERFACE BOARD
CM+9048614200
NETZTEIL
POWER SUPPLY
ALIMENTATION AVEC INTERRUPTEUR
CM+9042663040
TR, DECKELSCHALTER
CASSETTE UNIT SAFETY SWITCH
PORTE, INTERRUPTEUR DE COUVERCLE
CM+9517585851
NETZTEILKABEL
POWER SUPPLY CABLE
CABLE D'ALIMENTATION
CM+9517591950
FIREWIRE-STECKER MONTIERT
FIREWIRE PLUG
FIREWIRE FICHE
=Assembly
D =Standard part. For order please refer to the separate spare parts list for standard parts "DD+DIS011.93M"
* =No wearing part.
COVER
DOCUMENT CONTROL NOTE:
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 3
01-2009
CR 30-X
TYPE 5175/100/110
Chapter 5 / 7
DD+DIS325.06M
01
02
03,04
08
05
06
07
5175_CHAP05_02.CDR
SCAN UNIT
DOCUMENT CONTROL NOTE:
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 3
01-2009
CR 30-X
TYPE 5175/100/110
Chapter 5 / 8
DD+DIS325.06M
Pos. Nr. Teile-Nr.
Item no. Part No.
Pos. no. Nr. de Ref.
Benennung
Description
Denomination
CM+9517521802
PMI-BOARD CR30-X
PMI BOARD CR30-X
PMI BOARD CR30-X
CM+9517049600
CM+9047711070
CM+9047710970
CM+9517590104
PMT + LICHTSAMMLER
PMT + LIGHT COLLECTOR
PMT + CONDUIT DE LUMIRE
CM+9517524321
CM+9517527500
OPTIKBAUGRUPPE CR30-X
OPTIC MODULE CR30-X
ENSEMBLE OPTIQUE CR30-X
CM+9517585150
=Assembly
D =Standard part. For order please refer to the separate spare parts list for standard parts "DD+DIS011.93M"
* =No wearing part.
SCAN UNIT
DOCUMENT CONTROL NOTE:
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 3
01-2009
CR 30-X
TYPE 5175/100/110
Chapter 5 / 9
DD+DIS325.06M
02
03
04
05
06
5175_CHAP05_03.CDR
SLOWSCAN
DOCUMENT CONTROL NOTE:
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 3
01-2009
CR 30-X
TYPE 5175/100/110
Chapter 5 / 10
DD+DIS325.06M
Pos. Nr. Teile-Nr.
Item no. Part No.
Pos. no. Nr. de Ref.
Benennung
Description
Denomination
CM+9517524000
SLOWSCAN KOMPLETT
SLOWSCAN COMPLETE
SLOWSCAN COMPLET
CM+9517521200
SLOWSCAN BOARD
SLOWSCAN BOARD
SLOWSCAN BOARD
CM+9517526501
CM+9517515600
DISTRIBUTOR BOARD
DISTRIBUTOR BOARD
DISTRIBUTOR BOARD
CM+9517590801
CM+9517524702
KALBRIERPLATTE VOLLST.
CALIBRATION BOARD
PLAQUE DE CALIBRAGE
=Assembly
D =Standard part. For order please refer to the separate spare parts list for standard parts "DD+DIS011.93M"
* =No wearing part.
SLOWSCAN
DOCUMENT CONTROL NOTE:
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 3
01-2009
CR 30-X
TYPE 5175/100/110
Chapter 5 / 11
DD+DIS325.06M
01
02
12
03
04 (M5)
03
06
06
05
11 (M6)
09
08
07 (M4)
10
5175_CHAP05_04.CDR
DRAWER
DOCUMENT CONTROL NOTE:
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 3
01-2009
CR 30-X
TYPE 5175/100/110
Chapter 5 / 12
DD+DIS325.06M
Pos. Nr. Teile-Nr.
Item no. Part No.
Pos. no. Nr. de Ref.
Benennung
Description
Denomination
CM+9517536001
LSCHEINHEIT
ERASURE UNIT
UNITE D'EFFACEMENT
CM+9517585901
CM+9517590302
WALZENPAARE
INPUT ROLLERS
PAIRE DE ROULEAUX
CM+9517563500
CM+9517561006
TRANSPORTSCHLITTEN
DRAWER UNIT
CHARIOT DE TRANSPORT
CM+9517561250
CM+9517586900
CM+9517560010
CM+9949969500
SENSOR IR-REFLEX
ROLLER IP SENSOR
PHOTOCELLULE DE REFLEXION
10
CM+9517563400
HEBEKULISSENLAGER (1 STCK.)
BEARING FOR CONNECTING LINK (1 PIECE)
PALIER POUR COULISSE (1 PIECE)
11
CM+9517586800
12
CM+9517598900
SEITENBLECH TRANSPORTEINHEIT
SIDE PLATE TRANSPORT UNIT
SIDE PLATE TRANSPORT UNIT
=Assembly
D =Standard part. For order please refer to the separate spare parts list for standard parts "DD+DIS011.93M"
* =No wearing part.
DRAWER
DOCUMENT CONTROL NOTE:
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 3
01-2009
CR 30-X
TYPE 5175/100/110
Chapter 5 / 13
DD+DIS325.06M
01
05 (M3)
02
04
03
08
09
07
10 (M1)
06
12,13
11
5175_CHAP05_05.CDR
Edition 1, Revision 3
01-2009
CR 30-X
TYPE 5175/100/110
Chapter 5 / 14
DD+DIS325.06M
Pos. Nr. Teile-Nr.
Item no. Part No.
Pos. no. Nr. de Ref.
Benennung
Description
Denomination
CM+9517552003
KASSETTENFFNER
CASSETTE OPENER UNIT
SYSTME D'OUVERTURE CASSETTE
CM+9516050060
ANTISTATIKBRSTE
ANTISTATIC BRUSH - (CASSETTE UNIT ENTRY)
BROSSE ANTISTATIQUE
CM+9048614170
CHIP READER
CHIP READER
CHIP READER
CM+9517590500
ZAHNRIEMEN F. KASSETTENEINHEIT
TOOTHED BELTS FOR CASSETTE UNIT
POULIE A COURROIE POUR MODULE CASSETTE
CM+9517545190
CM+9517590403
KASSETTENAUFNAHME
CASSETTE FIXATION UNIT
LONGEMENT DE CASSETTE
CM+9517590600
MICROSCHALTER
CASSETTE EDGE SENSOR
MICRORUPTEUR
CM+9517552302
IP HANDLING BOARD
IP HANDLING BOARD
IP HANDLING BOARD
CM+9517558301
10
CM+9517558100
11
CM+9517556002
KASSETTENENTRIEGLER
CASSETTE UNLOCKING UNIT
CASSETTE DEVERROUILLAGE
12
CM+9517514100
13
CM+9517514110
=Assembly
D =Standard part. For order please refer to the separate spare parts list for standard parts "DD+DIS011.93M"
* =No wearing part.
Edition 1, Revision 3
01-2009
CR 30-X
TYPE 5175/100/110
Chapter 5 / 15
DD+DIS325.06M
01
2x
2x
2 x 14 (DIN7)
(Mounting of gears in transport unit)
M2x10 (DIN912)
(Mounting of cassette edge sensor)
1x
2x
Spring washer D = 2,2 mm (DIN127)
(Mounting of cassette edge sensor)
Special screw M3
(Mounting of permanent magnets
at drawer unit)
3x
4x
Screw M 4 x 12 (DIN912)
(Mounting of covers)
Screw M 4 x 8 (DIN912)
(Mounting of cassette unit)
2x
3x
Spring washer D = 4,3 mm (DIN127)
(Mounting of cassette unit)
M2 (DIN 439)
(Mounting of cassette edge sensor)
Micro fuses
4x
2 x 0.5 A
2 x 2.0 A
3 x 3.5 A
2 x 7.0 A
02
03
5 x 1.0 A
3 x 2.5 A
2 x 5.0 A
04
06
05
5175_CHAP05_06.CDR
Edition 1, Revision 3
01-2009
CR 30-X
TYPE 5175/100/110
Chapter 5 / 16
DD+DIS325.06M
Pos. Nr. Teile-Nr.
Item no. Part No.
Pos. no. Nr. de Ref.
Benennung
Description
Denomination
CM+9517590701
CR30-X KLEINTEILSET
CR30-X SMALL PARTS KIT
CR30-X KIT PICES DTACHES
CM+9517591401
CM+9517591600
CM+9517509331
CM+9517591301
CM+9517590900
TYPVERPACKUNG CR30-X
TYPE PACKAGING CR30-X
EMBALLAGE TYPE CR30-X
=Assembly
D =Standard part. For order please refer to the separate spare parts list for standard parts "DD+DIS011.93M"
* =No wearing part.
Edition 1, Revision 3
01-2009
CR 30-X
TYPE 5175/100/110
Chapter 5 / 17
DD+DIS325.06M
REAR SIDE
05* - 09*
01* - 04*
5175_CHAP05_07.CDR
CASSETTE CR MD 4.0 T
DOCUMENT CONTROL NOTE:
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 3
01-2009
CR 30-X
TYPE 5175/100/110
Chapter 5 / 18
DD+DIS325.06M
Pos. Nr. Teile-Nr.
Item no. Part No.
Pos. no. Nr. de Ref.
Benennung
Description
Denomination
CM+9518052002
SCHUBLADE + IP 35X43
TRAY + IP 35X43
TIROIR + IP 35X43
CM+9518252002
SCHUBLADE + IP 24X30
TRAY + IP 24X30
TIROIR + IP 24X30
CM+9518352002
SCHUBLADE + IP 18X24
TRAY + IP 18X24
TIROIR + IP 18X24
CM+9518452002
SCHUBLADE + IP 15X30
TRAY + IP 15X30
TIROIR + IP 15X30
CM+9518053002
LEERKASSETTE 35X43
BLANK CASSETTE 35X43
CASSETTE BLANC 35X43
CM+9518253002
LEERKASSETTE 24X30
BLANK CASSETTE 24X30
CASSETTE BLANC 24X30
CM+9518353002
LEERKASSETTE 18X24
BLANK CASSETTE 18X24
CASSETTE BLANC 18X24
CM+9518453002
LEERKASSETTE 15X30
BLANK CASSETTE 15X30
CASSETTE BLANC 15X30
CM+9518012201
=Assembly
D =Standard part. For order please refer to the separate spare parts list for standard parts "DD+DIS011.93M"
* =No wearing part.
CASSETTE CR MD 4.0 T
DOCUMENT CONTROL NOTE:
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 3
01-2009
CR 30-X
TYPE 5175/100/110
Chapter 5 / 19
DD+DIS325.06M
'Repair'
'Installation'
number of machines
extension of the service area
local service structure (centralized / decentralized)
stockpiling
M 'Maintenance'
'Local stock'
version
not applicable
Edition 1, Revision 3
01-2009
CR 30-X
Type 5175/0100/0110
Chapter 5 / 20
DD+DIS325.06M
Type Overview
This spare parts list is valid for the following machine type(s):
Specification
Device Name
Type No.
CR 30-X
5175/0100
Digitizer (WhiteLabel)
5175/0110
Accessory Overview
Following accesssories are separately available:
Order number
Accessory
TABLE FOR CR30-X
EWRP3
E2M3H
CR MOBILE KIT
EBYO1
Edition 1, Revision 3
01-2009
CR 30-X
Type 5175/0100/0110
Chapter 5 / 21
Chapter 8
HEALTHCARE
Manufacturing Standard
Modifications
Imaging Services
CR 30-X
Type 5175 /100/110
Document History
Edition.
Revision
Release
Date
Changes:
1.0
05-2008
Initial Release
Referenced Documents
Document
Title
n.a.
n.a.
Edition 1, Revision 0
05-2008 printed in Germany
DD+DIS147.08E
Manufacturer
Agfa HealthCare N.V.
Publisher
Agfa-Gevaert HealthCare GmbH
Tegernseer Landstrae 161
D - 81539 Mnchen
Germany
Copyright 2008 Agfa HealthCare N.V.
All rights reserved.
Technical modifications reserved.
AGFA and the Agfa-Rhombus are trademarks of Agfa HealthCare N.V.
WARNING:
Improper operation or service activities may cause damage or injuries.
INSTRUCTION:
(1)
(2)
Strictly observe all safety directions within the "Generic Safety Directions" and on
the product.
NOTE:
To verify the latest version of single documents and of Service Manuals refer to the
Document Type Order List in the GSO library.
Edition 1, Revision 0
05-2008
CR 30-X
Type 5175 /100/110
Chapter 8 / 2
DD+DIS147.08E
OVERVIEW OF MODIFICATIONS:
Timeline
Modification
Referenced Documents
2008
April Introduction Software CRE_1509
2007
November Introduction Software CRE_1405
Chapter 1, Controls,
Connectors and Setup
Procedure, DD+DIS150.06E
Rev. 1
Edition 1, Revision 0
05-2008
CR 30-X
Type 5175 /100/110
Chapter 8 / 3
DD+DIS147.08E
Timeline
Modification
Referenced Documents
2006
December Introduction of Software Patch CRE_1206C
October Added Post Scan Brush
Edition 1, Revision 0
05-2008
CR 30-X
Type 5175 /100/110
Chapter 8 / 4
DD+DIS147.08E
Product(s) &
Serial
Number(s)
Part
Number(s)
Not available via spare part channel, only via GSO library.
Referenced
Document(s)
Product(s) &
Serial
Number(s)
Part
Number(s)
Not available via spare part channel, only via GSO library.
Referenced
Document(s)
Product(s) &
Serial
Number(s)
Part
Number(s)
Referenced
Document(s)
Edition 1, Revision 0
05-2008
CR 30-X
Type 5175 /100/110
Chapter 8 / 5
DD+DIS147.08E
Product(s) &
Serial
Number(s)
Part
Number(s)
Referenced
Document(s)
SN 1610
Product(s) &
Serial
Number(s)
Part
Number(s)
Referenced
Document(s)
SN 1633
Product(s) &
Serial
Number(s)
Part
Number(s)
Referenced
Document(s)
Edition 1, Revision 0
05-2008
CR 30-X
Type 5175 /100/110
Chapter 8 / 6
DD+DIS147.08E
Product(s) &
Serial
Number(s)
Part
Number(s)
Referenced
Document(s)
SN 1532
Product(s) &
Serial
Number(s)
Part
Number(s)
Referenced
Document(s)
Product(s) &
Serial
Number(s)
Part
Number(s)
Referenced
Document(s)
Edition 1, Revision 0
05-2008
CR 30-X
Type 5175 /100/110
Chapter 8 / 7
DD+DIS147.08E
10
11
Product(s) &
Serial
Number(s)
Part
Number(s)
Referenced
Document(s)
SN 1701
Product(s) &
Serial
Number(s)
Part
Number(s)
Referenced
Document(s)
Edition 1, Revision 0
05-2008
CR 30-X
Type 5175 /100/110
Chapter 8 / 8
DD+DIS147.08E
12
13
Product(s) &
Serial
Number(s)
Part
Number(s)
Referenced
Document(s)
To have small parts kit (Screws, nuts, fuses) available in the field.
Product(s) &
Serial
Number(s)
Part
Number(s)
Referenced
Document(s)
Edition 1, Revision 0
05-2008
CR 30-X
Type 5175 /100/110
Chapter 8 / 9
DD+DIS147.08E
14
15
Product(s) &
Serial
Number(s)
Part
Number(s)
CM+8.5175.2470.2
Referenced
Document(s)
Product(s) &
Serial
Number(s)
Part
Number(s)
Edition 1, Revision 0
05-2008
CR 30-X
Type 5175 /100/110
Chapter 8 / 10
DD+DIS147.08E
16
17
18
Product(s) &
Serial
Number(s)
Part
Number(s)
Not available via spare part channel, only via GSO library.
Referenced
Document(s)
Product(s) &
Serial
Number(s)
Part
Number(s)
Referenced
Document(s)
SN 2301
Product(s) &
Serial
Number(s)
SN 2730
SN 6001
Part
Number(s)
Not available via spare part channel, only via GSO library.
Referenced
Document(s)
Edition 1, Revision 0
05-2008
CR 30-X
Type 5175 /100/110
Chapter 8 / 11
DD+DIS147.08E
19
20
Product(s) &
Serial
Number(s)
Part
Number(s)
Referenced
Document(s)
New functions:
Product(s) &
Serial
Number(s)
SN 3626
SN 6132
Part
Number(s)
Not available via spare part channel, only via GSO library.
Referenced
Document(s)
Edition 1, Revision 0
05-2008
CR 30-X
Type 5175 /100/110
Chapter 8 / 12
HEALTHCARE
Chapter 9
Imaging Services
Maintenance
CR 30-X
Type 5175 / 100/110
Edition 1, Revision 1
01-2009 printed in Germany
DD+DIS171.06E
Maintenance
Manufacturer
Agfa HealthCare N.V.
Publisher
Agfa-Gevaert HealthCare GmbH
Tegernseer Landstrae 161
D - 81539 Mnchen
Germany
Copyright 2008 Agfa HealthCare N.V.
All rights reserved.
Technical modifications reserved.
AGFA and the Agfa-Rhombus are trademarks of Agfa HealthCare N.V.
WARNING:
Improper operation or service activities may cause damage or injuries.
INSTRUCTION:
(1)
(2)
Strictly observe all safety directions within the "Generic Safety Directions" and on
the product.
NOTE:
To verify the latest version of single documents and of Service Manuals refer to the
Document Type Order List in the GSO library.
Edition 1, Revision 1
01-2009
CR 30-X
Type 5175 / 100/110
Chapter 9 / 2
DD+DIS171.06E
Maintenance
Document History
Edition.
Revision
Release
Date
Changes
compared to previous Version 1.0
1.1
01-2009
Referenced Documents
Document
Title
DD+DIS150.06E
Edition 1, Revision 1
01-2009
CR 30-X
Type 5175 / 100/110
Chapter 9 / 3
DD+DIS171.06E
Maintenance
LIST OF CONTENTS
1.2
Required Time...........................................................................................................................6
1.3
Required Tools..........................................................................................................................7
1.4
1.5
1.6
DIAGNOSTICS .........................................................................................................................9
2.1
2.2
2.3
2.4
2.5
2.6
3.2
3.3
3.4
Checking the Teflon Tape and Cassette Fixation Steel Band at the Cassette Unit ...............17
3.5
3.6
3.7
3.8
4.2
4.3
4.4
Edition 1, Revision 1
01-2009
CR 30-X
Type 5175 / 100/110
Chapter 9 / 4
DD+DIS171.06E
Maintenance
5.2
5.3
Edition 1, Revision 1
01-2009
CR 30-X
Type 5175 / 100/110
Chapter 9 / 5
DD+DIS171.06E
Maintenance
1.1
Maintenance Frequency
The maintenance has to be carried out:
1.2
once a year
Required Time
REQUIRED TIME:
approximately 2 hours
Edition 1, Revision 1
01-2009
CR 30-X
Type 5175 / 100/110
Chapter 9 / 6
DD+DIS171.06E
1.3
Maintenance
Required Tools
Description
Hex key 3 mm
1.4
1.5
Description
CM+9.9999.9288.0 or
sales order code EQP6D
Canned Air
commercially available
Lint-free cloth
commercially available
* The last digit in the spare part number indicates the spare part revision at release of this
document. When ordering, the actual revision of the spare part is delivered.
Edition 1, Revision 1
01-2009
CR 30-X
Type 5175 / 100/110
Chapter 9 / 7
DD+DIS171.06E
1.6
Maintenance
* The last digit in the spare part number indicates the spare part revision at release of
this document. When ordering, the actual revision of the spare part is delivered.
NOTE:
Replacing the parts mentioned above is required in exceptional cases only.
For details refer to sections 3.1 and 3.4.
Edition 1, Revision 1
01-2009
CR 30-X
Type 5175 / 100/110
Chapter 9 / 8
DD+DIS171.06E
Maintenance
Diagnostics
2.1
2.2
2.3
Ask the customer for any problem that appeared since the last maintenance.
(2)
Figure 1
2.4
In the CR30-X Service Client select General Device Info Info Counter".
(2)
(3)
(4)
(5)
(6)
Click on the errors displayed in the Lists tab to check the timestamp: This
shows, when the error occurred.
(7)
Check, whether one of these errors is sporadic and can be solved via software
upgrade, cleaning or a replacement.
Edition 1, Revision 1
01-2009
CR 30-X
Type 5175 / 100/110
Chapter 9 / 9
DD+DIS171.06E
2.5
Maintenance
(2)
(3)
Select all events (select upper most event, then select lowest event with the shift
key pressed).
(4)
(5)
Figure 2
Edition 1, Revision 1
01-2009
CR 30-X
Type 5175 / 100/110
Chapter 9 / 10
DD+DIS171.06E
2.6
Maintenance
Cu filter
(1)
NOTE:
FLFS cassettes can be recognized by a
sticker on the cassettes.
Genrad Cassette
FLFS Cassette
FLFS
FLFS
Figure 3
(2)
(3)
(4)
(5)
517509ag.cdr
Figure 4
(6)
Edition 1, Revision 1
01-2009
CR 30-X
Type 5175 / 100/110
Chapter 9 / 11
DD+DIS171.06E
Maintenance
IMPORTANT:
After cleaning, wait for about 10 minutes to allow the solvents to evaporate.
(7)
(8)
Figure 6
(9) Mount the Cu filter at the modality.
(10) Place the cassette in parallel to the
rotating anode.
Rotating
Anode
X-Ray tube
Cathode
517509aa.cdr
NOT OK
OK
Figure 7
(11) Expose the cassette with 10 Gy.
NOTE:
Select the following exposure parameters using a 1.5 mm Cu filter to obtain
a dose of 10 Gy:
12 mAs
75 kVp
large focus.
The values are approximate and may vary among the X-ray devices.
(12) Turn the cassette by 180.
(13) Again expose with 10 Gy.
(14) Remove the CU filter from the modality.
DOCUMENT CONTROL NOTE:
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 1
01-2009
CR 30-X
Type 5175 / 100/110
Chapter 9 / 12
DD+DIS171.06E
Maintenance
24 cm
Slow scan 35 cm
direction
24 cm
15 cm
18 cm
517509ab.cdr
Evaluation:
Symptom
Stripes in slow scan
direction
Possible reason
Dust on the light
collector
Spot artifacts
Procedure
Cleaning of the scan line is required.
This is done during step 3.6 Cleaning of the
Scan Line.
sibly the scan rollers have to be cleaned. This
is checked during step 3.7,
Cleaning of the lower Scan Rollers.
(20) Store the cassette aside: It is used once more for image quality check after
maintenance.
(21) Keep the image for comparison after maintenance.
Edition 1, Revision 1
01-2009
CR 30-X
Type 5175 / 100/110
Chapter 9 / 13
DD+DIS171.06E
Maintenance
3.1
Visual Inspection
REQUIRED TOOLS:
Flashlight
Hex key 3 mm
(1)
(2)
(3)
(4)
(5)
(6)
(7)
(9)
Edition 1, Revision 1
01-2009
CR 30-X
Type 5175 / 100/110
Chapter 9 / 14
DD+DIS171.06E
Maintenance
IMPORTANT:
For digitizers which still have the old side
plate with straight edge installed:
It is recommended to replace the side plate
by the Side Plate Transport Unit (left hand
side) preventively.
Spare part number see section 1.6.
Installation instructions are part of delivery
of the spare part.
Figure 10:
Cassette side plate with sloped rear edge
(new design)
(10) Remove screw at erasure unit.
(11) Disconnect two cables at erasure
unit.
(12) Take out erasure unit.
Figure 11
CAUTION:
IP jams possible if IP guide plate is deformed.
Do not bend IP guide plate at removal.
(13) Slide IP guide plate to the left and lift
it out.
(14) Check condition of the 4 belts in the
cassette unit.
Figure 12
DOCUMENT CONTROL NOTE:
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 1
01-2009
CR 30-X
Type 5175 / 100/110
Chapter 9 / 15
DD+DIS171.06E
3.2
Maintenance
Cleaning General
REQUIRED TOOLS:
Vacuum cleaner
(1)
Figure 13
3.3
Figure 14
Edition 1, Revision 1
01-2009
CR 30-X
Type 5175 / 100/110
Chapter 9 / 16
DD+DIS171.06E
3.4
Maintenance
Checking the Teflon Tape and Cassette Fixation Steel Band at the
Cassette Unit
Note:
The Teflon tape is only installed in digitizers manufactured till October 2008.
(1) Check the condition of the Teflon tape and the cassette fixation steel band in the
cassette unit:
The cassette fixation steel band may not show any bend or sticking part coming
from the Teflon tape.
Replace the Teflon tape by the slide bar, which is part of delivery of the
Cassette Fixation Steel Band, spare part number: CM+9 5175 5830 1
For details see instructions which are part of delivery of the spare part.
Teflon Tape
Edition 1, Revision 1
01-2009
CR 30-X
Type 5175 / 100/110
Chapter 9 / 17
DD+DIS171.06E
(2)
Maintenance
Figure 17
NOTE:
The slide bar is part of delivery of the
Cassette Fixation Steel Band, spare part
number see section 1.6.
It is installed in digitizers manufactured as
of October 2008. It can easily be
recognized by 2 clamps at the left side of
the cassette unit.
Figure 18
3.5
Figure 19
DOCUMENT CONTROL NOTE:
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 1
01-2009
CR 30-X
Type 5175 / 100/110
Chapter 9 / 18
DD+DIS171.06E
3.6
Maintenance
Figure 20
Edition 1, Revision 1
01-2009
CR 30-X
Type 5175 / 100/110
Chapter 9 / 19
DD+DIS171.06E
3.7
Maintenance
IMPORTANT:
Do not clean the upper rollers: They do not require any cleaning.
(1)
(2)
(3)
(4)
Edition 1, Revision 1
01-2009
CR 30-X
Type 5175 / 100/110
Chapter 9 / 20
DD+DIS171.06E
Maintenance
(5)
(6)
3.8
(7)
Figure 22
Re-mount the guide plates at the pre-scan and posts scan side.
(8)
(9)
(2)
Keep damaged or defective cassettes for the customer conversation at the end of
the preventive maintenance.
Maintenance Power on
4.1
(2)
(3)
(4)
(5)
(6)
(7)
Edition 1, Revision 1
01-2009
CR 30-X
Type 5175 / 100/110
Chapter 9 / 21
DD+DIS171.06E
Maintenance
Figure 23
4.2
Pull out the cassette unit approx. 1 cm (approx. 0.4 in.): The "cassette unit open
safety switch" has to switch off the digitizer.
If the test fails, the switch has to be adjusted or it is defective.
For adjustment instructions refer to Technical Documentation chapter 3.6.
For replacement instructions refer to Technical Documentation chapter 3.5.
(2)
Edition 1, Revision 1
01-2009
CR 30-X
Type 5175 / 100/110
Chapter 9 / 22
DD+DIS171.06E
4.3
Maintenance
(2)
Figure 24
4.4
4.4.1
(2)
(3)
Rotating
Anode
X-Ray tube
Cathode
517509aa.cdr
NOT OK
OK
Figure 25
DOCUMENT CONTROL NOTE:
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 1
01-2009
CR 30-X
Type 5175 / 100/110
Chapter 9 / 23
DD+DIS171.06E
Maintenance
(4)
NOTE:
Select the following exposure parameters using a 1.5 mm Cu filter
(CM+9 5155 1015 2) to obtain a dose of 10 Gy:
12 mAs
75 kVp
large focus
The values are approximate and may vary among the X-ray devices.
(5)
(6)
(7)
(8)
(9)
IMPORTANT:
By selection of examination type "System Diagnosis - Flatfield", the correct settings
for image processing of the flatfield are used. Any other examination type may give a
result which cannot be compared with the limit pattern.
(10)
(11)
Edition 1, Revision 1
01-2009
CR 30-X
Type 5175 / 100/110
Chapter 9 / 24
DD+DIS171.06E
4.4.2
Maintenance
Compare the flatfield made at the beginning of the maintenance procedure with
the one made under step 4.4.1.
Evaluation:
(2)
If the image made during step 4.4.1 visually looks as good or better than the
flatfield made at the beginning of the maintenance procedure, the image
quality is okay.
If there are unacceptable effects in slow scan or fast scan direction, refer to
the troubleshooting guide, DD+DIS151.08E.
Check the image quality of the flatfield for absence of any stripes and large area
inhomogeneities.
4.4.3
If the flatfield image shows stripes in fast scan or slow scan direction or large
area inhomogeneities compare the image with the limit pattern.
See section 4.4.3.
Comparing the Flatfield with the Test Patterns Lowest Acceptable Quality
IMPORTANT:
This step is only required, if the flatfield image made under step 4.4.1 shows stripes in
fast scan or slow scan direction or large area inhomogeneities.
(1)
Insert the limit pattern CD-ROM (part of delivery) into the CD-ROM Drive.
(2)
(3)
(4)
(5)
(6)
Select OK.
(7)
(8)
(9)
(10) At the lightbox or the viewing station visually compare the scanned flatfield with
the limit pattern. The limit pattern represents the lowest acceptable image quality
in slow scan and fast scan direction: Compare the stripes in slow scan and in
fast scan direction.
See also IMPORTANT note on page 27 if a FLFS cassette is used for flatfield
evaluation.
DOCUMENT CONTROL NOTE:
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 1
01-2009
CR 30-X
Type 5175 / 100/110
Chapter 9 / 25
DD+DIS171.06E
Maintenance
Figure 26
Evaluation:
If the scanned image visually looks as good or better than the limit pattern,
the image quality is okay.
If there are unacceptable effects in slow scan or fast scan direction, see the
troubleshooting guide, DD+DIS151.08E.
NOTE:
The slow scan direction is always the long side of the image plate.
30 cm
43 cm
30 cm
15 cm
24 cm
Slow scan 35 cm
direction
24 cm
18 cm
517509ab.cdr
Edition 1, Revision 1
01-2009
CR 30-X
Type 5175 / 100/110
Chapter 9 / 26
DD+DIS171.06E
Maintenance
IMPORTANT:
If you use a FLFS cassette for flatfield exposure, the image will be slightly darker in
the upper and lower image area where the backscatter protection is removed (approx.
1 cm each). These darker zones also have to be used for image quality evaluation in
slow scan and fast scan direction.
Flatfield with Genrad Cassette
Effect exaggerated
Genrad Cassette
FLFS Cassette
Backscatter
Protection
FLFS
10 mm
10 mm
FLFS
Figure 28
Edition 1, Revision 1
01-2009
CR 30-X
Type 5175 / 100/110
Chapter 9 / 27
DD+DIS171.06E
Maintenance
5.1
(1)
(2)
Figure 29
DOCUMENT CONTROL NOTE:
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 1
01-2009
CR 30-X
Type 5175 / 100/110
Chapter 9 / 28
DD+DIS171.06E
5.2
Maintenance
(2)
(3)
(4)
(5)
Save the file with date and serial number in the file name.
(6)
Copy the file additionally to a USB stick and store the file on a save place.
5.3
Customer Conversation
(1)
Fill out the Maintenance checklist and hand it out to the customer.
(2)
Edition 1, Revision 1
01-2009
CR 30-X
Type 5175 / 100/110
Chapter 9 / 29
DD+DIS171.06E
Maintenance
This page is intentionally left blank to enable print out of even and odd pages on
duplex printers.
Edition 1, Revision 1
01-2009
CR 30-X
Type 5175 / 100/110
Chapter 9 / 30
Maintenance
Maintenance Checklist
DD+DIS171.06E
Maintenance Checklist
Work Instruction for order no.
SN
Film cycles
NOTE:
Maintenance must be carried out according to the maintenance instructions
DD+DIS171.06E
Diagnostics
Done
Questioning of the
customer
Checking images on
the workstation
Error statistics
Infocounter analysis
Checking current
Image Quality
Edition 1, Revision 1
01-2009
CR 30-X
Type 5175 / 100/110
Chapter 9 / 31
Maintenance
Maintenance Checklist
DD+DIS171.06E
Done
Visual check of
o Power and FireWire cable
o Covers
o Cables inside the digitizer, especially at cassette
unit left hand side
o 4 belts in the cassette unit
Cleaning; General
Cleaning of the
Erasure Unit
Cleaning of scan
rollers
Checking of Cassette
Condition
Done
Performing stall
calibration for all
motors.
Performing Test
Cycles
Technical Image
Quality Check of the
System
Edition 1, Revision 1
01-2009
CR 30-X
Type 5175 / 100/110
Chapter 9 / 32
Maintenance
Maintenance Checklist
DD+DIS171.06E
Completion
Done
Reset of Maintenance
Indicator
Creating a Backup of
Device specific Data
Creating a Backup
Storage on hard disk and on service PC
Remarks:
..........................................................................................................................................
..........................................................................................................................................
..........................................................................................................................................
..........................................................................................................................................
..........................................................................................................................................
..........................................................................................................................................
..........................................................................................................................................
..........................................................................................................................................
.............................................................
Date / Signature Service Technician
.........................................................
Date / Signature Customer
Edition 1, Revision 1
01-2009
CR 30-X
Type 5175 / 100/110
Chapter 9 / 33
HEALTHCARE
Chapter 10
Imaging Services
Service Bulletins
CR 30-X
Type 5175 /100/110
Please note
Service Bulletins are not part of the Service Manual for Download.
Please download the Service Bulletins from the GSO Library:
MEDNET GSO Library Computed Radiography > CR Digitizers > CR30-X
Service Bulletin
Edition 1, Revision 0
04-2008 printed in Germany
DD+DIS150.06E
Service Bulletins
Edition 1, Revision 0
04-2008
CR 30-X
Type 5175 /100/110
Chapter 10 / 2
HEALTHCARE
Chapter 11
Imaging Services
Installation Planning
CR 30-X
Type 5175 / 100/110
Edition 1, Revision 3
Edition 1, Revision 3
07-2008 printed in Germany
DD+DIS172.06E
Installation Planning
Manufacturer
Agfa HealthCare N.V.
Publisher
Agfa-Gevaert HealthCare GmbH
Tegernseer Landstrae 161
D - 81539 Mnchen
Germany
Copyright 2008 Agfa HealthCare N.V.
All rights reserved.
Technical modifications reserved.
AGFA and the Agfa-Rhombus are trademarks of Agfa HealthCare N.V.
WARNING:
Improper operation or service activities may cause damage or injuries.
INSTRUCTION:
(1)
(2)
Strictly observe all safety directions within the "Generic Safety Directions" and on
the product.
NOTE:
To verify the latest version of single documents and of Service Manuals refer to the
Document Type Order List in the GSO library.
Edition 1, Revision 3
07-2008
CR 30-X
Type 5175 / 100/110
Chapter 11 / 3
DD+DIS172.06E
Installation Planning
Document History
Edition.
Revision
Release
Date
1.3
07-2008
Changes
compared to previous Version 1.2:
Edition 1, Revision 3
07-2008
CR 30-X
Type 5175 / 100/110
Chapter 11 / 4
DD+DIS172.06E
Installation Planning
LIST OF CONTENTS
Accessories.............................................................................................................................10
7.1
During Operation.....................................................................................................................18
7.2
8.2
SPECIFICATIONS..................................................................................................................22
9.1
9.2
9.3
9.4
9.5
9.6
9.7
9.8
Interface Data..........................................................................................................................24
9.9
Service Data............................................................................................................................24
9.10
9.11
9.11.1
9.11.2
Edition 1, Revision 3
07-2008
CR 30-X
Type 5175 / 100/110
Chapter 11 / 5
DD+DIS172.06E
10
Installation Planning
9.11.3
9.11.4
Weight .....................................................................................................................................25
9.11.5
SAFETY STANDARDS...........................................................................................................26
Edition 1, Revision 3
07-2008
CR 30-X
Type 5175 / 100/110
Chapter 11 / 6
DD+DIS172.06E
Installation Planning
Installation Planning-Checklist
Step
Goal
See
Okay
section
1.
Ask the Agfa sales representative To schedule an installation date with the
for a contact person in the hospital customer.
2.
3.
4.
3.1
5.
6.
7.
8.
9.
10
Last
Page
Edition 1, Revision 3
07-2008
CR 30-X
Type 5175 / 100/110
Chapter 11 / 7
DD+DIS172.06E
Installation Planning
Product Description
This is the Installation Planning Document for the CR 30-X digitizers type 5175,
subtypes 100 and 110.
Name of the digitizers:
Subtype 100:
CR 30-X
Subtype 110:
2.1
System Overview
The digitizer is part of the digitizer system, consisting of:
Digitizer
NX Workstation (Tower or laptop version)
Cassette with Image Plate (IP)
Digitizer
NX Workstation
Tower Version
NX Workstation
Laptop Version
Digitizer
Cassette
Cassette
517511ea.cdr
Figure 1
NOTE:
This document describes the installation planning of the digitizer.
For more information to the NX processing station and optional printer refer to
MEDNET, GSO library.
DOCUMENT CONTROL NOTE:
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 3
07-2008
CR 30-X
Type 5175 / 100/110
Chapter 11 / 8
DD+DIS172.06E
Installation Planning
part number
name
F2.5175.1030.
F7.0470.3400.
F8.5155.1015.
F7.0477.1101.
F7.0477.1097.
F7.0345.0024.
F7.5175.1043.2
F7.5175.0049.5
F8.5175.1075.3
10
F8.5175.1065.
11
12
13
F8.5175.0933.
F8.5175.0940.
F8.5175.9070.
14
F7.5175.1350.
qty.
Verpackung Beipack
Packaging accessories
Anschluleitung 3ADR I-2500-S (Europa)
Power cable 3ADR I-2500-S (Europe)
Cu-Filter verpackt
Cu-filter packed
Netzanschlultg. HG 125V/10A
Power cable HG 125V/10A (USA)
Firewire-Kabel 6P/m -6P/m 4,5m
Firewire-cable 6P/m -6P/m 4,5m
Installations-Report DD+DIS018.06E
Installation Report DD+DIS018.06E
CE Declaration
CE Declaration of Conformity
Packliste Beipack CR 30-X
Packing list accessories CR 30-X
CD CR 30-X System Benutzerhandbuch
CD CR 30-X System User Documentation
Installationsanleitung DD+DIS317.07E
Installation Instruction DD+DIS317.07E
Digitizer Software CD CRE_1509
Software CR30 Pattern CD 1.0
CR30-X Small Parts Kit
RoHS-Beiblatt CR 30-X
RoHS sheet
ok
1
1
1
1
1
1
1
1
1
1
1
1
1
1
Abstandshalter / Spacer
F2.1098.0795.
F7.0374.0065.
1
1
F7.5175.1546.
D7.0213.816.24
Schraube / Screw
D6.0096.420.00
Scheibe / Washer
L7.0371.7036.
Steckschelle / Clamp
packed by: ..
5
4
3A
3
2
1A
1
39814
39925
39978
39802
39795
39754
39714
2.07.08
18.03.08
30.01.08
04.10.07
14.08.07
04.06.07
15.05.07
nd.Zust.
Kommt
vor
MitteilungsNr.
Datum
AGLCH
Datum
AGPMK Ersteller
AGRMU Prfer
AGPSN
AGPMK
AGPSN
AGPSN
Name
23.04.07
23.04.07
Name
AGPSN
AGPSN
ab
Agfa-Gevaert
HealthCare GmbH
Packing list
CR30-X
Accessories Agfa
Benennung:
Typ:
5175/100
Sachnummer:
F7.5175.0049.5
Ersatz fr:
Blatt von
Blatt: 1/1
Edition 1, Revision 3
07-2008
CR 30-X
Type 5175 / 100/110
Chapter 11 / 9
DD+DIS172.06E
Installation Planning
IMPORTANT:
NOTE:
3.1
The digitizer is delivered with all parts, which are required to connect it to a
NX workstation.
Accessories
Item
Order Number
EWRP3
All-in-one cart
E2M3H
UPS Powerware
5115,
120 V Version
EGPSE
UPS Powerware
5115,
220 V Version
EGPTG
EWRQ5
Comment
For dimensions refer to 4.1
Can be used for digitizer and NX workstation. For
technical specifications see sections 8, 8.2 and 9.4.
See safety note below.
517511bb.cdr
Remote Service
Not applicable
Base Plate
Edition 1, Revision 3
07-2008
CR 30-X
Type 5175 / 100/110
Chapter 11 / 10
DD+DIS172.06E
Installation Planning
Warning:
Images can be lost due to power failure.
Connect the equipment to an un-interruptible power supply (UPS) or an institutional
standby generator.
NOTE:
Previously a CR Inroom X-Ray Shielding was available as accessory.
This is not available anymore. Take appropriate measures to protect the digitizer
and cassettes with image plates from scattered radiation. See also section 7.1.
NOTE to Accessory "Remote Service":
In general it is possible to diagnose and repair the digitizer to a certain extent
remotely.
To enable remote service of the digitizer remote access has first to be organized.
Remote Access is established via the Agfa own Secure Remote Service System
(SRSS).
For more information to the SRSS see MEDNET GSO Library - Technical
Services - Service Delivery - SRSS.
Edition 1, Revision 3
07-2008
CR 30-X
Type 5175 / 100/110
Chapter 11 / 11
DD+DIS172.06E
Installation Planning
Digitizer Dimensions
Dimensions in cm [inch].
Digitizer closed:
49.7 [19.6]
517511aa.cdr
78.6 [30.9]
69.3 [27.3]
Figure 3
Digitizer opened:
49.7 [19.6]
517511ab.cdr
69.3 [27.3]
100.6 [39.6]
Figure 4
Edition 1, Revision 3
07-2008
CR 30-X
Type 5175 / 100/110
Chapter 11 / 12
DD+DIS172.06E
4.1
Installation Planning
All-in-one cart
IMPORTANT:
If the customer uses a different table, observe the stability of the table: It must be
able to carry the weight of 98 kg (216.1 lb) and may not be subject to excessive
shock and vibrations from other sources. See also corresponding safety note in
section 6.
Dimensions see next page.
Edition 1, Revision 3
07-2008
CR 30-X
Type 5175 / 100/110
Chapter 11 / 13
DD+DIS172.06E
TOP
LEFT
66,5 cm [26,2"]
FRONT
73,4 cm [28,9"]
Installation Planning
65,7 cm [25,7"]
FRONT
LEFT
75,7 cm [29,8"]
155 cm [61,1"]
77,0 cm [30,3"]
70,0 cm [27,6"]
TOP
Edition 1, Revision 3
07-2008
CR 30-X
Type 5175 / 100/110
Chapter 11 / 14
DD+DIS172.06E
Installation Planning
> 10 [3.9] *
> 50 [19.7]
**
60 [23.6]
60 [23.6]
80 [31.5]
80 [31.5]
60 [23.6]
60 [23.6]
Figure 7
NOTE:
If installed on the optional wheel table (see section 4.1) the digitizer can also be
installed in a corner.
Be aware, that the mains switch is in the rear. See section 8.1. At a corner installation
the mains switch may not be accessible.
IMPORTANT:
*The clearance in the rear for electrical connectors and to provide sufficient air flow
has to be at least 10 cm [3.9 in] in any case.
**Also at least 50 cm [19.7 in] space has to be provided on the right hand side to
allow cleaning of the scan line with the claning brush. This can be performed by the
customer.
Edition 1, Revision 3
07-2008
CR 30-X
Type 5175 / 100/110
Chapter 11 / 15
DD+DIS172.06E
Installation Planning
System Integration
NOTE:
*Fast ID Workflow means: The cassette is not identified before exposure, but waits
for patient data entry at the NX after the cassette is entered into the digitizer. Only
when the patient data are entered at the NX and sent to the digitizer the scan cycle
begins.
NX Workstation
Tower Version
Digitizer
Cassette
75 cm [29.5 in]
recommended
PC
517511ai.cdr
6 pin to 4 pin
FWI converter
(part of laptop delivery)
Figure 8
Edition 1, Revision 3
07-2008
CR 30-X
Type 5175 / 100/110
Chapter 11 / 16
DD+DIS172.06E
Installation Planning
WARNING:
Using a FWI cable > 4.5 m may lead to unstable or no communication
between digitizer and NX workstation.
Only use the FireWire cable which is part of delivery. Longer cables are not
supported.
WARNING:
When different combinations of equipment are used in various medical
environments a potential difference (V) can exist between the protective
earths in different localities. If the protective earthing fails this potential
difference can cause a HAZARD for the OPERATOR or for the PATIENT.
To comply with ISO 60601-1 (annex I) all computers and peripherals must be
connected to the same power source.
WARNING:
Excessive vibrations during scanning may decrease image quality.
The structure and stability of the used table need to be suitable in relation with the
size and weight of the system. The table may not be subject to excessive shock
and vibrations from other sources.
Edition 1, Revision 3
07-2008
CR 30-X
Type 5175 / 100/110
Chapter 11 / 17
DD+DIS172.06E
Installation Planning
7.1
During Operation
min.:
max.:
best:
Ambient temperature:
min.:
(at 25C (77F) ambient temperature) max.:
best:
Relative humidity:
+15C (59F)
+30C (86F)
+20C (68F) to 25 C (77F)
15%
80%
30% to 60%
Radiation influences:
Warning:
Image plate is sensitive for X-rays. Poor image quality possible.
The digitizer and the cassette storage shall be protected against X-ray radiation this
way, that the annual dose equivalent at the installation place will not exceed 1 mSv.
7.2
Ambient Light:
Transport
Edition 1, Revision 3
07-2008
CR 30-X
Type 5175 / 100/110
Chapter 11 / 18
DD+DIS172.06E
Installation Planning
Electrical Connections
Warning:
Electrical device. Shock possible.
INSTRUCTION:
Installation
guidelines
UL:
Mains connection
in the installation
room
VDE:
UL:
Mains cable
(part of delivery)
Euro-Version:
US/CAN-Version:
IMPORTANT:
The digitizer is delivered with two power cables for
US/Europe. If you need a different cable, organize it
locally.
Edition 1, Revision 3
07-2008
CR 30-X
Type 5175 / 100/110
Chapter 11 / 19
DD+DIS172.06E
Installation Planning
517511aj.cdr
4
to wall outlet
USB
to wall outlet
517501ra.cdr
Figure 9
* concerning digitizer mains cable delivery see IMPORTANT note
previous page.
Edition 1, Revision 3
07-2008
CR 30-X
Type 5175 / 100/110
Chapter 11 / 20
DD+DIS172.06E
8.1
Installation Planning
Mains switch
517511ad.cdr
100 - 240 V / 50 - 60 Hz AC
power supply input
Intergrounding
Connector
FireWire
Connector (IEEE1394)
Figure 10
8.2
Four 10 A, IEC-320
Receptacles
10 A, IEC-320
Input Connector
Four 5 - 15
Receptacles
517511aj.cdr
Figure 11
Edition 1, Revision 3
07-2008
CR 30-X
Type 5175 / 100/110
Chapter 11 / 21
DD+DIS172.06E
Specifications
9.1
Type Overview
9.2
Installation Planning
Type
ABC Code
Serial Numbers
5175/100
ETDPS
1001
5175/110
EXF1O
6001
Functional Data
Throughput 35x43 cm cassettes
Pixels per line x scan lines
15 x 30 cm
18 x 24 cm
24 x 30 cm
35 x 43 cm
9.3
100-240 AC (autosensing)
50/60
120 standby
320 peak
Edition 1, Revision 3
07-2008
CR 30-X
Type 5175 / 100/110
Chapter 11 / 22
DD+DIS172.06E
9.4
Installation Planning
120 = Default
230 = Default
Climatic data
Heat emission [kJ/h]
432 standby
1152 maximum
9.7
50/60 (autosensing)
>5
750
192 x 150 x 270 / 7.6 x 5.9 x
10.6
7.8 / 17.2
refer to www.powerware.com
Weight [kg/lb]
More info
9.6
Frequency [Hz]
Battery runtime [min.]
Output Power [W]
Dimensions (HxWxD) [mm/in]
9.5
in standby:
in operation:
5 dB(A) maximum
65 dB(A) maximum
60
Edition 1, Revision 3
07-2008
CR 30-X
Type 5175 / 100/110
Chapter 11 / 23
DD+DIS172.06E
9.8
Installation Planning
Service Data
Infocounter
Error diagnostics built in
Remote / preventive service capable
9.10
Interface Data
Interface to NX workstation
9.9
yes
yes
yes
Within 1 hour
After 2 hours
After 24 hours
MD40
Recommended readout time
70% of the stored energy is still
present with no visible loss of
information upon readout
Image retention still exceeds
45%. Slightly reduced image
quality
Edition 1, Revision 3
07-2008
CR 30-X
Type 5175 / 100/110
Chapter 11 / 24
DD+DIS172.06E
Installation Planning
9.11
9.11.1
9.11.2
78.6 / 30.9
69.3 / 27.3
49.7 / 19.6
Weight
Digitizer alone (kg / lb)
Digitizer with packing incl. palette
(kg / lb)
9.11.5
90 / 35.4
80 / 31.5
69 / 27.2
98 / 216.1
124 / 273.4
Dimensions Pallet
Length / depth (cm / inch)
Width (cm / inch)
Height (cm / inch)
90 / 35.4
80 / 31.5
15 / 6
Edition 1, Revision 3
07-2008
CR 30-X
Type 5175 / 100/110
Chapter 11 / 25
DD+DIS172.06E
10
Installation Planning
Safety Standards
The DIGITIZER complies with:
See also section 10 "Compliance" in the appended chapter "Generic Safety Directions"
The device complies with the EN 60601-1, 2006 standard for Information Technology.
This means that, although it is absolutely safe, patients may not come in direct contact
with the equipment. Therefore the operator console must be placed outside a radius of
1.5 m around the patient.
R = 1.5m
1.5m
1.5m
Patient environment
Edition 1, Revision 3
07-2008
CR 30-X
Type 5175 / 100/110
Chapter 11 / 26
Installation Planning
Checklist
DD+DIS172.06E
In order to guarantee an effective and smooth installation process within the targeted time frame,
the following tasks of the checklist below must be carried out prior to the installation of the device.
Check and discuss all the required measures for the installation by means of this checklist.
Remarks on the individual items can be noted on the back of the list.
After completion please hand out this checklist to your local Agfa representative
Name of Agfa FSE:
Client name:
______________________________________________________
Client address:
_________________________________________________________
_________________________________________________________
Contact person,
name and title:
_________________________________________________________
______________________
Extension ____________________
Task
OK
Edition 1, Revision 3
07-2008
CR 30-X
Type 5175 / 100/110
Chapter 11 / 27
DD+DIS150.06E
Chapter 12
List of Contents
Glossary ........................................................................................1
Edition 1, Revision 0
CR30-X
Type 5175/100
Chapter 12 / I
Agfa Company Confidential
DD+DIS150.06E
Glossary
Term
Short description
Gy
ADC
AE title
Application Entity
ASAP - Package
BOL
Begin Of Line
BOS
Begin Of Scan
Browser
Calibration
CAN bus
CAS
CCM
Collimation
CPF file
CPU
CR
Computed Radiography
CU-Filter
DICOM
Edition 1, Revision 0
CR30-X
Type 5175/100
Chapter 12 / 1
Agfa Company Confidential
DD+DIS150.06E
Term
Short description
Dosimeter
DR
Direct Radiography
DRA
Dynamic range
FireWire
FSE
GenRad
General Radiology
Grid
GSC
GUI
Heel Effect
HIS
Histogram
ID Chip
ID Station
IDentification Station
Infocounter
IP
IP
Chapter 12 / 2
Agfa Company Confidential
CR30-X
Type 5175/100
Edition 1, Revision 0
DD+DIS150.06E
Term
Short description
IP address
Jitter
Laser
LCD
LED
Leeds Phantom
LUT
Look Up Table
mR
MCBF
MTTI
MTTR
MUSICA
Node
Pixel
RAM
RIS
ROI
Region Of Interest
ROM
RS232 interface
Edition 1, Revision 0
CR30-X
Type 5175/100
Chapter 12 / 3
Agfa Company Confidential
DD+DIS150.06E
Term
Short description
Shading calibration
Slow-scan direction
Throughput
Twain Interface
UI
UPS
URL
UTP - cable
W/L
Chapter 12 / 4
Agfa Company Confidential
CR30-X
Type 5175/100
Edition 1, Revision 0
Order no.:
DD+DIS150.06E
CR30-X
Type 5175/100
1st Edition
HEALTHCARE
ReadMeFirst
Service Manual for Download
Imaging Services
print.
add comments.
Document History
Edition.
Revision
Release
Date
Changes
compared to previous Version 1.1
1.2
06-2008
Edition 1, Revision 2
06-2008 printed in Germany
Agfa Company Confidential
ReadMeFirst
IMPORTANT:
Preferably print this manual double-sided:
This PDF manual contains empty pages at the end of several chapters, to have the
next chapter starting with an uneven page number when printed doubles-sided.
If printed one-sided, dispose these empty pages.
Preferably print circuit diagrams on a DIN A3 or ANSI B (Ledger) printer, if
available.
Some pages especially circuit diagrams for equipment have been created on paper
size larger than DIN A4/Letter. Printing these pages on DIN A4/Letter may result in
reduced legibility. It is recommended to print these pages separately on a DIN A3 or
ANSI B (Ledger) printer.
Edition 1, Revision 2
06-2008
Page 2 of 5
Agfa Company Confidential
ReadMeFirst
2.1
(2)
Write down or remember the shown PDF page number. See Figure 1.
(4)
Select "Print".
(5)
(6)
(7)
Select OK.
Edition 1, Revision 2
06-2008
Page 3 of 5
Agfa Company Confidential
ReadMeFirst
NOTE:
Shrink to printable area may be named on other Adobe Reader versions
reduce to printer margins or shrink oversized pages to paper size.
2.2
Select "Print".
(2)
Select All.
(3)
(4)
Select OK.
Adding Comments
If you open this file in an Adobe Reader version 7, the comment toolbar will
show-up.
This allows adding comments, to highlight or underline text and many more text
manipulations.
Edition 1, Revision 2
06-2008
Page 4 of 5
Agfa Company Confidential
ReadMeFirst
3.1
(1)
In the drop down menu Comment & Markup select "Show comments List".
3.2
(2)
Select the desired comments: Press the CTRL-key for multiple selections.
(3)
(4)
Importing Comments
(1)
In the drop down menu Comment & Markup select "Show comments List".
(2)
(3)
NOTE:
The imported comments possibly appear on different pages, if the file where the
comments have been imported has a different number of pages.
Edition 1, Revision 2
06-2008
Page 5 of 5
Agfa Company Confidential