MT Self-Study Note
MT Self-Study Note
MT Self-Study Note
Study Note
MT
Leo.liu
Content
I.
II.
III.
IV.
V.
VI.
VII.
Introduction
Physics
Equipment & Materials
Testing practices
Process control
Example indications
Quizzes
I. Introduction
1.
2.
3.
Introduction
Basic of concepts
History of MPI
1. Introduction
Magnetic particle inspection (MPI) is a nondestructive testing method used for defect
detection. MPI is fast and relatively easy to apply, and part surface preparation is not as
critical as it is for some other NDT methods. These characteristics make MPI one of the most
widely utilized nondestructive testing methods.
MPI uses magnetic fields and small magnetic particles (i.e.iron filings) to detect flaws in
components. The only requirement from an inspectability standpoint is that the component
being inspected must be made of a ferromagnetic material such as iron, nickel, cobalt, or
some of their alloys. Ferromagnetic materials are materials that can be magnetized to a level
that will allow the inspection to be effective.
The method is used to inspect a variety of product forms including castings, forgings, and
weldments. Many different industries use magnetic particle inspection for determining a
component's fitness-for-use. Some examples of industries that use magnetic particle
inspection are the structural steel, automotive, petrochemical, power generation, and
aerospace industries. Underwater inspection is another area where magnetic particle
inspection may be used to test items such as offshore structures and underwater pipelines.
2. Basic of concepts
In theory, magnetic particle inspection (MPI) is a relatively simple concept. It can be
considered as a combination of two nondestructive testing methods: magnetic flux leakage
testing and visual testing. Consider the case of a bar magnet. It has a magnetic field in and
around the magnet. Any place that a magnetic line of force exits or enters the magnet is
called a pole. A pole where a magnetic line of force exits the magnet is called a north pole
and a pole where a line of force enters the magnet is called a south pole.
NS
2. Basic of concepts
When a bar magnet is broken in the center of its length, two complete bar magnets with
magnetic poles on each end of each piece will result. If the magnet is just cracked but not
broken completely in two, a north and south pole will form at each edge of the crack. The
magnetic field exits the north pole and reenters at the south pole. The magnetic field spreads
out when it encounters the small air gap created by the crack because the air cannot support
as much magnetic field per unit volume as the magnet can. When the field spreads out, it
appears to leak out of the material and, thus is called a flux leakage field.
2. Basic of concepts
If iron particles are sprinkled on a cracked magnet, the particles will be attracted to and
cluster not only at the poles at the ends of the magnet, but also at the poles at the edges of
the crack. This cluster of particles is much easier to see than the actual crack and this is the
basis for magnetic particle inspection.
The first step in a magnetic particle inspection is to magnetize the component that is to be
inspected. If any defects on or near the surface are present, the defects will create a leakage
field. After the component has been magnetized, iron particles, either in a dry or wet
suspended form, are applied to the surface of the magnetized part. The particles will be
attracted and cluster at the flux leakage fields, thus forming a visible indication that the
inspector can detect.
3. History of MPI
https://www.ndeed.org/EducationResources/CommunityCollege/MagParticle/Introduction/history.php
1928 Electyro-Magnetic Steel Testing Device (MPI) made by the Equipment and Engineering
Company Ltd. (ECO) of Strand, England.
II. Physics
1.
2.
3.
4.
5.
6.
7.
8.
9.
10.
Magnetism
Magnetic Mat'ls
Magnetic Domains
Magnetic Fields
Electromag. Fields
Field From a Coil
Mag Properties
Hysteresis Loop
Permeability
Field Orientation
11.
12.
13.
14.
15.
16.
Magnetization of Mat'ls
Magnetizing Current
Longitudinal Mag Fields
Circular Mag Fields
Demagnetization
Measuring Mag Fields
1. Magnetism
Magnets are very common items in the workplace and household. Uses of magnets range from
holding pictures on the refrigerator to causing torque in electric motors. Most people are familiar
with the general properties of magnets but are less familiar with the source of magnetism. The
traditional concept of magnetism centers around the magnetic field and what is know as a dipole.
The term "magnetic field" simply describes a volume of space where there is a change in energy
within that volume. This change in energy can be detected and measured. The location where a
magnetic field can be detected exiting or entering a material is called a magnetic pole. Magnetic
poles have never been detected in isolation but always occur in pairs, hence the name dipole.
Therefore, a dipole is an object that has a magnetic pole on one end and a second, equal but
opposite, magnetic pole on the other.
A bar magnet can be considered a dipole with a north pole at one end and south pole at the other.
A magnetic field can be measured leaving the dipole at the north pole and returning the magnet at
the south pole. If a magnet is cut in two, two magnets or dipoles are created out of one. This
sectioning and creation of dipoles can continue to the atomic level. Therefore, the source of
magnetism lies in the basic building block of all matter...the atom.
1. Magnetism
The Source of Magnetism
All matter is composed of atoms, and atoms are composed of protons, neutrons and electrons. The
protons and neutrons are located in the atoms nucleus and the electrons are in constant motion
around the nucleus. Electrons carry a negative electrical charge() and produce a magnetic field
as they move through space. A magnetic field is produced whenever an electrical charge is in
motion. The strength of this field is called the magnetic moment().
This may be hard to visualize on a subatomic scale but consider electric current flowing through a
conductor. When the electrons (electric current) are flowing through the conductor, a magnetic
field forms around the conductor. The magnetic field can be detected using a compass. The
magnetic field will place a force on the compass needle, which is another example of a dipole.
Since all matter is comprised of atoms, all materials are affected in some way by a magnetic field.
However, not all materials react the same way. This will be explored more in the next section.
2. Magnetic Materials
Diamagnetic, Paramagnetic, and Ferromagnetic Materials
When a material is placed within a magnetic field, the magnetic forces of the materials electrons
will be affected. This effect is known as Faradays Law of Magnetic Induction. However, materials
can react quite differently to the presence of an external magnetic field. This reaction is dependent
on a number of factors, such as the atomic and molecular() structure of the material, and the
net magnetic field associated with the atoms. The magnetic moments associated with atoms have
three origins. These are the electron motion, the change in motion caused by an external
magnetic field, and the spin of the electrons.
In most atoms, electrons occur in pairs. Electrons in a pair spin in opposite directions. So, when
electrons are paired together, their opposite spins cause their magnetic fields to cancel each other.
Therefore, no net magnetic field exists. Alternately, materials with some unpaired electrons will
have a net magnetic field and will react more to an external field. Most materials can be classified
as diamagnetic, paramagnetic or ferromagnetic.
2. Magnetic Materials
Diamagnetic
Diamagnetic materials have a weak, negative susceptibility to magnetic fields. Diamagnetic
materials are slightly repelled by a magnetic field and the material does not retain the magnetic
properties when the external field is removed. In diamagnetic materials all the electron are
paired so there is no permanent net magnetic moment per atom. Diamagnetic properties arise
from the realignment of the electron paths under the influence of an external magnetic field. Most
elements in the periodic table, including copper, silver, and gold, are diamagnetic.
Paramagnetic
Paramagnetic materials have a small, positive susceptibility to magnetic fields. These materials are
slightly attracted by a magnetic field and the material does not retain the magnetic properties
when the external field is removed. Paramagnetic properties are due to the presence of some
unpaired electrons, and from the realignment of the electron paths caused by the external
magnetic field. Paramagnetic materials include magnesium, molybdenum, lithium, and tantalum.
2. Magnetic Materials
Ferromagnetic
Ferromagnetic materials have a large, positive susceptibility to an external magnetic field. They
exhibit a strong attraction to magnetic fields and are able to retain their magnetic properties
after the external field has been removed. Ferromagnetic materials have some unpaired electrons
so their atoms have a net magnetic moment. They get their strong magnetic properties due to the
presence of magnetic domains(). In these domains, large numbers of atom's moments (1012 to
1015) are aligned parallel so that the magnetic force within the domain is strong. When a
ferromagnetic material is in the unmagnitized state, the domains are nearly randomly organized
and the net magnetic field for the part as a whole is zero. When a magnetizing force is applied, the
domains become aligned to produce a strong magnetic field within the part. Iron, nickel, and
cobalt are examples of ferromagnetic materials. Components with these materials are commonly
inspected using the magnetic particle method.
2. Magnetic Materials
3. Magnetic Domains
Ferromagnetic materials get their magnetic properties not only because their atoms carry a
magnetic moment but also because the material is made up of small regions known as magnetic
domains. In each domain, all of the atomic dipoles are coupled together in a preferential direction.
This alignment develops as the material develops its crystalline structure during solidification from
the molten state. Magnetic domains can be detected using Magnetic Force Microscopy (MFM) and
images of the domains like the one shown below can be constructed.
3. Magnetic Domains
During solidification, a trillion or more atom moments are aligned parallel so that the magnetic
force within the domain is strong in one direction. Ferromagnetic materials are said to be
characterized by "spontaneous magnetization" since they obtain saturation magnetization in each
of the domains without an external magnetic field being applied. Even though the domains are
magnetically saturated, the bulk material may not show any signs of magnetism because the
domains develop themselves and are randomly oriented relative to each other.
Ferromagnetic materials become magnetized when the magnetic domains within the material are
aligned. This can be done by placing the material in a strong external magnetic field or by passing
electrical current through the material. Some or all of the domains can become aligned. The more
domains that are aligned, the stronger the magnetic field in the material. When all of the domains
are aligned, the material is said to be magnetically saturated. When a material is magnetically
saturated, no additional amount of external magnetization force will cause an increase in its
internal level of magnetization.
4. Magnetic Fields
Magnetic Field In and Around a Bar Magnet
As discussed previously, a magnetic field is a change in energy within a volume of space. The
magnetic field surrounding a bar magnet can be seen in the magnetograph below. A magnetograph
can be created by placing a piece of paper over a magnet and sprinkling the paper with iron filings.
The particles align themselves with the lines of magnetic force produced by the magnet. The
magnetic lines of force show where the magnetic field exits the material at one pole and reenters
the material at another pole along the length of the magnet. It should be noted that the magnetic
lines of force exist in three dimensions but are only seen in two dimensions in the image.
It can be seen in the magnetograph that there are poles all along the length of the magnet but that
the poles are concentrated at the ends of the magnet. The area where the exit poles are
concentrated is called the magnet's north pole and the area where the entrance poles are
concentrated is called the magnet's south pole.
4. Magnetic Fields
Magnetic Fields in and around Horseshoe and Ring Magnets
Magnets come in a variety of shapes and one of the more common is the horseshoe (U) magnet.
The horseshoe magnet has north and south poles just like a bar magnet but the magnet is curved
so the poles lie in the same plane. The magnetic lines of force flow from pole to pole just like in the
bar magnet. However, since the poles are located closer together and a more direct path exists for
the lines of flux to travel between the poles, the magnetic field is concentrated between the poles.
If a bar magnet was placed across the end of a horseshoe magnet or if a magnet was formed in the
shape of a ring, the lines of magnetic force would not even need to enter the air. The value of such
a magnet where the magnetic field is completely contained with the material probably has limited
use. However, it is important to understand that the magnetic field can flow in loop within a
material. (See section on circular magnetism for more information).
4. Magnetic Fields
General Properties of Magnetic Lines of Force
Magnetic lines of force have a number of important properties, which include:
They seek the path of least resistance between opposite magnetic poles. In a single bar magnet
as shown to the right, they attempt to form closed loops from pole to pole.
They never cross one another.
They all have the same strength.
Their density decreases (they spread out) when they move from an area of higher
permeability() to an area of lower permeability.
Their density decreases with increasing distance from the poles.
They are considered to have direction as if flowing, though no actual movement occurs.
They flow from the south pole to the north pole within a material and north pole to south pole
in air.
5. Electromagnetic Fields
Magnets are not the only source of magnetic fields. In 1820, Hans Christian Oersted()
discovered that an electric current flowing through a wire caused a nearby compass to deflect. This
indicated that the current in the wire was generating a magnetic field. Oersted studied the nature
of the magnetic field around the long straight wire. He found that the magnetic field existed in
circular form around the wire and that the intensity of the field was directly proportional to the
amount of current carried by the wire. He also found that the strength of the field was strongest
next to the wire and diminished with distance from the conductor until it could no longer be
detected. In most conductors, the magnetic field exists only as long as the current is flowing (i.e. an
electrical charge is in motion). However, in ferromagnetic materials the electric current will cause
some or all of the magnetic domains to align and a residual magnetic field will remain.
5. Electromagnetic Fields
Oersted also noticed that the direction of the magnetic field was dependent on the direction of the
electrical current in the wire. A three-dimensional representation of the magnetic field is shown
below. There is a simple rule for remembering the direction of the magnetic field around a
conductor. It is called the right-hand clasp rule. If a person grasps a conductor in one's right
hand with the thumb pointing in the direction of the current, the fingers will circle the conductor in
the direction of the magnetic field.
5. Electromagnetic Fields
A word of caution about the right-hand clasp rule
For the right-hand rule to work, one important thing that must be remembered about the
direction of current flow. Standard convention has current flowing from the positive terminal to
the negative terminal. This convention is credited to Benjamin Franklin who theorized that electric
current was due to a positive charge moving from the positive terminal to the negative terminal.
However, it was later discovered that it is the movement of the negatively charged electron that is
responsible for electrical current. Rather than changing several centuries of theory and equations,
Franklin's convention is still used today.
SI Units
(Sommerfeld)
Quantity
SI Units
(Kennelly)
CGS Units
(Gaussian)
Field
A/m
A/m
oersteds
Flux Density
(Magnetic Induction)
tesla
tesla
gauss
Flux
weber
weber
maxwell
Magnetization
A/m
erg/Oe-cm
The number of magnetic lines of force cutting through a plane of a given area at a right angle is
known as the magnetic flux density, B. The flux density or magnetic induction has the tesla as its
unit. One tesla is equal to 1 Newton/(A/m). From these units, it can be seen that the flux density is
a measure of the force applied to a particle by the magnetic field. The Gauss is the CGS unit for flux
density and is commonly used by US industry. One gauss represents one line of flux passing
through one square centimeter of air oriented 90 degrees to the flux flow.
BTesla1T=1N/(A/m)Gauss
11T=10,000Gauss
9. Permeability
As previously mentioned, permeability (m) is a material property that describes the ease with
which a magnetic flux is established in a component. It is the ratio of the flux density (B) created
within a material to the magnetizing field (H) and is represented by the following equation:
m = B/H
It is clear that this equation describes the slope of the curve at any point on the hysteresis loop.
The permeability value given in papers and reference materials is usually the maximum
permeability or the maximum relative permeability. The maximum permeability is the point where
the slope of the B/H curve for the unmagnetized material is the greatest. This point is often taken
as the point where a straight line from the origin is tangent to the B/H curve.
9. Permeability
The relative permeability() is arrived at by taking the ratio of the material's
permeability to the permeability in free space (air).
9. Permeability
The shape of the hysteresis loop tells a great deal about the material being magnetized. The
hysteresis curves of two different materials are shown in the graph.
Relative to other materials, a material with a wider hysteresis loop has:
Lower Permeability()
Higher Retentivity()
Higher Coercivity()
Higher Reluctance()
Higher Residual Magnetism()
Relative to other materials, a material with the narrower hysteresis loop has:
Higher Permeability
Lower Retentivity
Lower Coercivity
Lower Reluctance
Lower Residual Magnetism
9. Permeability
In magnetic particle testing, the level of residual magnetism is important. Residual magnetic fields
are affected by the permeability, which can be related to the carbon content and alloying of the
material. A component with high carbon content will have low permeability and will retain more
magnetic flux than a material with low carbon content.
MT
In the two B-H loops above, which one would indicative of a low carbon steel??
Blue one.
Watch this short movie showing the effect of field direction on indication visibility. (2.10)
Direct Current
Alternating Current
Rectified Alternating Current
Half Wave Rectified Alternating Current (HWAC)
Full Wave Rectified Alternating Current (FWAC) (Single Phase)
Three Phase Full Wave Rectified Alternating Current
Alternating Current
Alternating current (AC) reverses in direction at a rate of 50 or 60 cycles per second. In the United
States, 60 cycle current is the commercial norm but 50 cycle current is common in many countries.
Since AC is readily available in most facilities, it is convenient to make use of it for magnetic particle
inspection. However, when AC is used to induce a magnetic field in ferromagnetic materials, the
magnetic field will be limited to narrow region at the surface of the component. This phenomenon
is known as the "skin effect" and occurs because the changing magnetic field generates eddy
currents in the test object. The eddy currents produce a magnetic field that opposes the primary
field, thus reducing the net magnetic flux below the surface. Therefore, it is recommended that AC
be used only when the inspection is limited to surface defects.
ACDC
The magnetic field travels through the component from end to end with some flux loss along its
length as shown in the image to the right. Keep in mind that the magnetic lines of flux occur in
three dimensions and are only shown in 2D in the image. The magnetic lines of flux are much more
dense inside the ferromagnetic material than in air because ferromagnetic materials have much
higher permeability than does air. When the concentrated flux within the material comes to the air
at the end of the component, it must spread out since the air can not support as many lines of flux
per unit volume. To keep from crossing as they spread out, some of the magnetic lines of flux are
forced out the side of the component.
The field strength varies from zero at the center of the component to a maximum at the surface.
The field strength at the surface of the conductor decreases as the radius of the conductor
increases when the current strength is held constant. (However, a larger conductor is capable of
carrying more current.)
The field strength outside the conductor is directly proportional to the current strength. Inside
the conductor, the field strength is dependent on the current strength, magnetic permeability
of the material, and if magnetic, the location on the B-H curve.
The field strength outside the conductor decreases with distance from the conductor.
DCACDC
AC
2-292-30
MT
15. Demagnetization
After conducting a magnetic particle inspection, it is usually necessary to demagnetize the
component. Remanent magnetic fields can:
Removal of a field may be accomplished in several ways. This random orientation of the magnetic
domains can be achieved most effectively by heating the material above its curie temperature. The
curie temperature for a low carbon steel is 770 or 1390oF. When steel is heated above its curie
temperature, it will become austenitic and loses its magnetic properties. When it is cooled back
down, it will go through a reverse transformation and will contain no residual magnetic field. The
material should also be placed with it long axis in an east-west orientation to avoid any influence of
the Earth's magnetic field.
7701390oF
-
15. Demagnetization
It is often inconvenient to heat a material above its curie temperature to demagnetize it, so
another method that returns the material to a nearly unmagnetized state is commonly used.
Subjecting the component to a reversing and decreasing magnetic field will return the dipoles to a
nearly random orientation throughout the material. This can be accomplished by pulling a
component out and away from a coil with AC passing through it. The same can also be
accomplished using an electromagnetic yoke with AC selected. Also, many stationary magnetic
particle inspection units come with a demagnetization feature that slowly reduces the AC in a coil
in which the component is placed.
A field meter is often used to verify that the residual flux has been
removed from a component. Industry standards usually require
that the magnetic flux be reduced to less than 3 gauss after
completing a magnetic particle inspection.
3GMPI
Portable Equipment
Stationary Equipment
Lights
Field Strength Indicators
Magnetic Particles
Suspension Liquids
Permanent magnets
Electromagnets
Prods
Portable Coils and Conductive Cables
Portable Power Supplies
90
Watch these short movies showing a cable being used to establish magnetic fields in parts. Cable
wrapped around part (3.1.2mov). Cable wrapped through the part (3.1.3mov).
Wavelength Range
Class
UV-A 3,200~4,000 Angstroms
UV-B 2,800~3,200 Angstroms
UV-C 2,800~1,000 Angstroms
Class
Wavelength Range
UV-A 3,200~4,000 Angstroms
UV-B 2,800~3,200 Angstroms
UV-C 2,800~1,000 Angstroms
UV
400W10100W
micro-discharge light?
10
60,000uw/cm2 15in
QQI
QQI
QQI
QQI
Watch this short movie to see a Pie field gage in action (3.4).
5. Magnetic Particles
As mentioned previously, the particles that are used for magnetic particle inspection are a key
ingredient as they form the indications that alert the inspector to defects. Particles start out as tiny
milled (a machining process) pieces of iron or iron oxide. A pigment (somewhat like paint) is
bonded to their surfaces to give the particles color. The metal used for the particles has high
magnetic permeability and low retentivity. High magnetic permeability is important because it
makes the particles attract easily to small magnetic leakage fields from discontinuities, such as
flaws. Low retentivity is important because the particles themselves never become strongly
magnetized so they do not stick to each other or the surface of the part. Particles are available in a
dry mix or a wet solution.
5. Magnetic Particles
Dry Magnetic Particles
Dry magnetic particles can typically be purchased in red, black, gray, yellow and several other
colors so that a high level of contrast between the particles and the part being inspected can be
achieved. The size of the magnetic particles is also very important. Dry magnetic particle products
are produced to include a range of particle sizes. The fine particles are around 50 um (0.002 inch)
in size, and are about three times smaller in diameter and more than 20 times lighter than the
coarse particles (150 um or 0.006 inch). This make them more sensitive to the leakage fields from
very small discontinuities. However, dry testing particles cannot be made exclusively of the fine
particles. Coarser particles are needed to bridge large discontinuities and to reduce the powder's
dusty nature. Additionally, small particles easily adhere to surface contamination, such as remnant
dirt or moisture, and get trapped in surface roughness features. It should also be recognized that
finer particles will be more easily blown away by the wind; therefore, windy conditions can reduce
the sensitivity of an inspection. Also, reclaiming the dry particles is not recommended because
the small particles are less likely to be recaptured and the "once used" mix will result in less
sensitive inspections.
150um
50um20
5. Magnetic Particles
The particle shape is also important. Long, slender particles tend align themselves along the lines
of magnetic force. However, research has shown that if dry powder consists only of long, slender
particles, the application process would be less than desirable. Elongated particles come from the
dispenser in clumps and lack the ability to flow freely and form the desired "cloud" of particles
floating on the component. Therefore, globular particles are added that are shorter. The mix of
globular and elongated particles result in a dry powder that flows well and maintains good
sensitivity. Most dry particle mixes have particles with L/D ratios between one and two.
L/D L/D
5. Magnetic Particles
Wet Magnetic Particles
Magnetic particles are also supplied in a wet suspension such as water or oil. The wet magnetic
particle testing method is generally more sensitive than the dry because the suspension provides
the particles with more mobility and makes it possible for smaller particles to be used since dust
and adherence to surface contamination is reduced or eliminated. The wet method also makes it
easy to apply the particles uniformly to a relatively large area.
Wet method magnetic particles products differ from dry powder products in a number of ways.
One way is that both visible and fluorescent particles are available. Most nonfluorescent particles
are ferromagnetic iron oxides, which are either black or brown in color. Fluorescent particles are
coated with pigments that fluoresce when exposed to ultraviolet light. Particles that fluoresce
green-yellow are most common to take advantage of the peak color sensitivity of the eye but other
fluorescent colors are also available. (For more information on the color sensitivity of the eye...see
the material on penetrant inspection.)
5. Magnetic Particles
The particles used with the wet method are smaller in size than those used in the dry method for
the reasons mentioned above. The particles are typically 10 um (0.0004 inch) and smaller and the
synthetic iron oxides have particle diameters around 0.1 um (0.000004 inch). This very small size is
a result of the process used to form the particles and is not particularly desirable, as the particles
are almost too fine to settle out of suspension. However, due to their slight residual magnetism,
the oxide particles are present mostly in clusters that settle out of suspension much faster than the
individual particles. This makes it possible to see and measure the concentration of the particles
for process control purposes. Wet particles are also a mix of long slender and globular particles.
The carrier solutions can be water or oil-based. Water-based carriers form quicker indications, are
generally less expensive, present little or no fire hazard, give off no petrochemical fumes, and are
easier to clean from the part. Water-based solutions are usually formulated with a corrosion
inhibitor to offer some corrosion protection. However, oil-based carrier solutions offer superior
corrosion and hydrogen embrittlement protection to those materials that are prone to attack by
these mechanisms
6. Suspension Liquids
Suspension liquids used in the wet magnetic particle inspection method can be either a well
refined light petroleum distillate or water containing additives. Petroleum-based liquids are the
most desirable carriers because they provided good wetting of the surface of metallic parts.
However, water-based carriers are used more because of low cost, low fire hazard, and the ability
to form indications quicker than solvent-based carriers. Water-based carriers must contain wetting
agents to disrupt surface films of oil that may exist on the part and to aid in the dispersion of
magnetic particles in the carrier. The wetting agents create foaming as the solution is moved about,
so anti-foaming agents must be added. Also, since water promotes corrosion in ferrous materials,
corrosion inhibitors are usually added as well.
Petroleum based carriers are primarily used in systems where maintaining the proper particle
concentration is a concern. The petroleum based carriers require less maintenance because they
evaporate at a slower rate than the water-based carriers. Therefore, petroleum based carriers
might be a better choice for a system that gets only occasional use or when regularly adjusting the
carrier volume is undesirable. Modern solvent carriers are specifically designed with properties
that have flash points above 200oF and keep nocuous vapors low. Petroleum carriers are required
to meet certain specifications such as AMS 2641.
AMS2641
Dry Particles
Wet Suspension
Magnetic Rubber
Continuous & Residual Magnetization
Field Direction & Intensity
L/D Ratio
1. Dry Particles
In this magnetic particle testing technique, dry particles are dusted onto the surface of the test
object as the item is magnetized. Dry particle inspection is well suited for the inspections
conducted on rough surfaces. When an electromagnetic yoke is used, the AC or half wave DC
current creates a pulsating magnetic field that provides mobility to the powder. The primary
applications for dry powders are unground welds and rough as-cast surfaces.
ACDC
Dry particle inspection is also used to detect shallow subsurface cracks. Dry particles with half
wave DC is the best approach when inspecting for lack of root penetration in welds of thin
materials. Half wave DC with prods and dry particles is commonly used when inspecting large
castings for hot tears and cracks.
DC
1. Dry Particles
Steps in performing an inspection using dry particles
Prepare the part surface - the surface should be relatively clean but this is not as critical as it is
with liquid penetrant inspection. The surface must be free of grease, oil or other moisture that
could keep particles from moving freely. A thin layer of paint, rust or scale will reduce test
sensitivity but can sometimes be left in place with adequate results. Specifications often allow up
to 0.003 inch (0.076 mm) of a nonconductive coating (such as paint) and 0.001 inch max (0.025
mm) of a ferromagnetic coating (such as nickel) to be left on the surface. Any loose dirt, paint, rust
or scale must be removed.
Apply the magnetizing force - Use permanent magnets, an electromagnetic yoke, prods, a coil or
other means to establish the necessary magnetic flux.
Dust on the dry magnetic particles - Dust on a light layer of magnetic particles.
Gently blow off the excess powder - With the magnetizing force still applied, remove the excess
powder from the surface with a few gentle puffs of dry air. The force of the air needs to be strong
enough to remove the excess particles but not strong enough to dislodge particles held by a
magnetic flux leakage field.
Terminate the magnetizing force - If the magnetic flux is being generated with an electromagnet or
an electromagnetic field, the magnetizing force should be terminated. If permanent magnets are
being used, they can be left in place.
Inspect for indications - Look for areas where the magnetic particles are clustered.
VS
2G
performing an inspection on a standard specimen that is similar to the test component and has
known or artificial defects of the same type, size, and location as those expected in the test
component. QQI shims can sometimes be used as the artificial defects.
using a gauss meter with a Hall effect probe to measure the peak values of the tangent field at
the surface of the part in the region of interest. Most specifications call for a field strength of 30
to 60 gauss at the surface when the magnetizing force is applied.
looking for light furring at the ends of pipes or bars when performing dry particle inspections
on these and other uncomplicated shapes.
V. Process Control
1.
2.
3.
4.
5.
Particle Concentration
Suspension Contamination
Electrical system
Lighting
Eye Consideration
0.05ml1ml20
0.1ml1.5ml15
6030
0.1~0.4 ml1.2~2.4ml
2. Suspension Contamination
The suspension solution should also be examined for evidence of contamination. Contamination
primarily comes from inspected components. Oils, greases, sand, and dirt will be introduced to the
system through components. If the area is unusually dusty, the system will pickup dust or other
contaminates from the environment.
+
This examination is performed on the carrier and particles collected for concentration testing. The
graduated portion of the tube is viewed under ultraviolet and white light when fluorescent
particles are being used, and under white light when visible particles are being used. The magnetic
particles should be examined for foreign particles, such as dirt, paint chips and other solids.
Differences in color, layering or banding within the settled particles would indicate contamination.
Some contamination is to be expected but if the foreign matter exceeds 30 percent of the settled
solids, the solution should be replaced. 30%
2. Suspension Contamination
The liquid carrier portion of the solution should also be inspected for contamination. Oil in a water
bath and water in a solvent bath are the primary concerns. If the solution fluoresces brightly when
fluorescent particles are being used, this can be an indication that dye is being dislodged()
from the particles by the mixing pump. While not technically contamination, this condition should
be further evaluated by allowing the collected sample bath to set for 10 to 12 hours and viewed
under ultraviolet light. If a band that fluoresces brighter than the bulk of particles is evident on top
of the settled solids, the bath contains excessive unattached fluorescent pigments and should be
discarded.
10~12
2. Suspension Contamination
Water Break Test
A daily water break check is required to evaluate the surface wetting performance of water-based
carriers. The water break check simply involves flooding a clean surface similar to those being
inspected and observing the surface film. If a continuous film forms over the entire surface,
sufficient wetting agent is present. If the film of suspension breaks (water break) exposing the
surface of the component, insufficient wetting agent is present and the solution should be
adjusted or replaced.
Ammeter Check
It is important that the ammeter provide consistent and correct readings. If the meter is reading
low, over magnetization will occur and possibly result in excessive background "noise." If ammeter
readings are high, flux density could be too low to produce detectable indications. To verify
ammeter accuracy, a calibrated ammeter is connected in series with the output circuit and values
are compared to the equipment's ammeter values. Readings are taken at three output levels in the
working range. The equipment meter is not to deviate from the calibrated ammeter more than
10 percent or 50 amperes, whichever is greater. If the meter is found to be outside this range,
the condition must be corrected. It should be noted that when measuring half-wave rectified
current, the readings should be doubled.
10%50A
2
4. Lighting
Magnetic particle inspection predominately relies on visual inspection to detect any indications
that form. Therefore, lighting is a very important element of the inspection process. Obviously, the
lighting requirements are different for an inspection conducted using visible particles than they are
for an inspection conducted using fluorescent particles. The lighting requirements for each of these
techniques, as well as how light measurements are made, is discussed below.
4.1 Light Requirements When Using Visible Particles
Magnetic particle inspections that use visible particles can be conducted using natural or artificial
lighting. When using natural lighting, it is important to keep in mind that daylight varies from hour
to hour. Inspector must constantly stay aware of the lighting conditions and make adjustments
when needed. To improve the uniformity of lighting from one inspection to the next, the use of
artificial lighting is recommended. Artificial lighting should be white whenever possible and white
flood or halogen lamps are most commonly used. The light intensity is required to be 100 footcandles at the surface being inspected. It is advisable to choose a white light wattage that will
provide sufficient light, but avoid excessive reflected light that could distract from the inspection.
MT100fc 1000Lux
4. Lighting
4.2 Light Requirements When Using Fluorescent Particles
Ultraviolet Lighting
When performing a magnetic particle inspection using fluorescent particles, the condition of the
ultraviolet light and the ambient white light must be monitored. Standards and procedures require
verification of lens condition and light intensity. Black lights should never be used with a cracked
filter as the output of white light and harmful black light will be increased. The cleanliness of the
filter should also be checked regularly since a coating of solvent carrier, oil, or other foreign
material can reduce the intensity or light by as much as 50%. The filter should be checked visually
and cleaned as necessary before warming-up the light.
50%
4. Lighting
For UV lights used in component evaluations, the normally accepted intensity is 1000mwatts/cm2
when measured at 15 inches from the filter face (requirements can vary from 800 to
1200mwatts/cm2). The required check should be performed when a new bulb is installed, at
startup of the inspection cycle, if a change in intensity is noticed, or every eight hours if in
continuous use. Regularly checking the intensity of UV lights is very important because bulbs lose
intensity over time. In fact, a bulb that is near the end of its operating life will often have an
intensity of only 25% of its original output. Black light intensity will also be affected by voltage
variations, so it is important to provide constant voltage to the light. A bulb that produces
acceptable intensity at 120 volts will produce significantly less at 110 volts.
15inch1000/800~1200
lux
25%
4. Lighting
Ambient White Lighting
When performing a fluorescent magnetic particle inspection, it is important to keep white light to a
minimum as it will significantly reduce the inspectors ability to detect fluorescent indications. Light
levels of less than 2fc are required by most procedures and some procedures require it to be less
than 0.5fc at the inspection surface. Some specifications require that a white light intensity
measurement be made at 15 inches from a UV light source to verify that the white light is being
removed by the filter.
2fc20lux0.5fc
15ft
White Light for Indication Confirmation
While white light is held to a minimum in fluorescent inspections, procedures may require that
indications be evaluated under white light. The white light requirements for this evaluation are the
same as when performing an inspection with visible particles. The minimum light intensity at the
surface being inspected must be 100fc.
100fc
4. Lighting
>100fc1000lux
~1000mwarts/cm2 :
<2fc20lux
15ft:
4. Lighting
4.3 Light Measurement
Light intensity measurements are made using a radiometer(). A radiometer is an instrument
that translates light energy into an electrical current. Light striking a silicon photodiode(
) detector causes a charge to build up between internal layers. When an external circuit is
connected to the cell, an electrical current is produced. This current is linear with respect to
incident light. Some radiometers have the ability to measure both white and UV light, while others
require a separate sensor for each measurement. The sensing area should always be kept clean
and free of materials that could reduce or obstruct light reaching the sensor.
Radiometers are relatively unstable instruments and readings often change considerably over time.
Therefore, they must be calibrated regularly. They should be calibrated at least every six months. A
unit should be checked to make sure its calibration is current before taking any light readings.
6
Ultraviolet light measurements should be taken using a fixture to maintain a minimum distance of
15 inches from the filter face to the sensor. The sensor should be centered in the light field to
obtain the peak reading. UV spot lights are often focused which causes intensity readings to vary
considerable over a small area. White lights are seldom focused and, depending on the wattage,
will often produce in excess of the 100fc at 15 inches.
5. Eye Consideration
Eye Adaptation
Just as lighting is an important consideration in the inspection process, so is the eyes response to
light. Scientists have recently discovered that a special, tiny group of cells at the back of the eye
help tell the brain how much light there is, causing the pupil() to get bigger or smaller. The
change in pupil diameter is not instantaneous, therefore, eyes must be given time to adapt to
changing lighting conditions. When performing a fluorescent magnetic particle inspection, the eye
must be given time to adapt to the darkness of the inspection booth before beginning to look for
indications. Dark adaptation time of at least one minute is required by most procedures. Some
studies recommend adaptation time of five minutes if entering an inspection area from direct
sunlight. Inspectors should carefully adhere to the required adaptation time as it is quite easy to
overlook an indication when an inspection is started before the eyes have adjusted to the
darkened conditions.
1min
5min
5. Eye Consideration
Ocular Fluorescence
When ultraviolet light enters the human eye, the fluid that fills the eye fluoresces. This condition is
called ocular fluorescence, and while it is considered harmless, it is annoying and interferes with
vision while it exists. When working around ultraviolet lights, one should be careful not to look
directly into lights and to hold spot lights to avoid reflection. UV light will be reflected from
surfaces just as white light will, so it is advisable to consider placement of lights to avoid this
condition. Special filtered glasses may be worn by the inspector to remove all UV light from
reaching the eyes but allowing yellow-green light from fluorescent indications to pass. Technicians
should never wear darkened or photochromatic glasses as these glasses also filter or block light
from fluorescent indications.