ISO - TC - 213 - N - 658 - Vocabulario Del VIM
ISO - TC - 213 - N - 658 - Vocabulario Del VIM
ISO - TC - 213 - N - 658 - Vocabulario Del VIM
TMB
Voting begins on
2004-05-14
INTERNATIONAL ORGANIZATION FOR STANDARDIZATION
Voting terminates on
2004-09-14
In accordance with the ISO/IEC Directives, Part 1, 2001, Clause A.5, this draft is submitted to the ISO national
bodies for approval.
WARNING THIS DOCUMENT IS NOT AN ISO GUIDE. IT IS DISTRIBUTED FOR REVIEW AND COMMENT. IT IS SUBJECT TO CHANGE WITHOUT
NOTICE AND MAY NOT BE REFERRED TO AS A GUIDE.
RECIPIENTS OF THIS DRAFT ARE INVITED TO SUBMIT, WITH THEIR COMMENTS, NOTIFICATION OF ANY RELEVANT PATENT RIGHTS OF WHICH
THEY ARE AWARE AND TO PROVIDE SUPPORTING DOCUMENTATION.
IN ADDITION TO THEIR EVALUATION AS BEING ACCEPTABLE FOR INDUSTRIAL, TECHNOLOGICAL, COMMERCIAL AND USER PURPOSES,
DRAFT GUIDES MAY ON OCCASION HAVE TO BE CONSIDERED IN THE LIGHT OF THEIR POTENTIAL TO BECOME DOCUMENTS TO WHICH
REFERENCE MAY BE MADE IN NATIONAL REGULATIONS.
PDF disclaimer
This PDF file may contain embedded typefaces. In accordance with Adobe's licensing policy, this file may be printed or viewed but
shall not be edited unless the typefaces which are embedded are licensed to and installed on the computer performing the editing. In
downloading this file, parties accept therein the responsibility of not infringing Adobe's licensing policy. The ISO Central Secretariat
accepts no liability in this area.
Adobe is a trademark of Adobe Systems Incorporated.
Details of the software products used to create this PDF file can be found in the General Info relative to the file; the PDF-creation
parameters were optimized for printing. Every care has been taken to ensure that the file is suitable for use by ISO member bodies. In
the unlikely event that a problem relating to it is found, please inform the Central Secretariat at the address given below.
Copyright notice
This ISO document is a Draft Guide and is copyright-protected by ISO. Except as permitted under the
applicable laws of the user's country, neither this ISO draft nor any extract from it may be reproduced,
stored in a retrieval system or transmitted in any form or by any means, electronic, photocopying, recording
or otherwise, without prior written permission being secured.
Requests for permission to reproduce should be addressed to either ISO at the address below or ISO's
member body in the country of the requester.
ISO copyright office
Case postale 56 CH-1211 Geneva 20
Tel. + 41 22 749 01 11
Fax + 41 22 749 09 47
E-mail copyright@iso.org
Web www.iso.org
Reproduction may be subject to royalty payments or a licensing agreement.
Violators may be prosecuted.
ii
For convenience in the management of the enquiry ballot, this draft is submitted under the reference
ISO DGUIDE 99999. On publication the reference ISO DGUIDE 99999 will be discarded.
In accordance with the provisions of Council Resolution 15/1993 this document is circulated in the English
language only.
Conformment aux dispositions de la Rsolution du Conseil 15/1993, ce document est distribu en version
anglaise seulement.
To expedite distribution, this document is circulated as received from the committee secretariat. ISO Central
Secretariat work of editing and text composition will be undertaken at publication stage.
Pour acclrer la distribution, le prsent document est distribu tel qu'il est parvenu du secrtariat du comit.
Le travail de rdaction et de composition de texte sera effectu au Secrtariat central de l'ISO au stade de
publication.
iii
CONTENTS
Foreword.2
1 Quantities and unit..4
2 Measurement.. 13
3 Devices for measurements..26
4 Characteristics of measuring systems...30
5 Measurement standards, etalons36
Annex A: Concepts used in the classical approach (CA)43
to measurement
Informative Annex: Diagram concepts.47
Bibliography.... 48
List of acronyms..49
Alphabetic index..50
1.1
(1.1)
quantity
property of a phenomenon, body, or substance, to which a magnitude can be assigned
NOTES
1 The concept quantity can be subdivided in two levels, general concept and individual
concept.
EXAMPLES
length, l
radius, r
wavelength, O
energy,
kinetic energy, T
E
heat, Q
electric charge, Q
electric resistance, R
amount-of-substance concentration
of entity B, cB
number concentration of entity B,
CB
Rockwell C hardness, HRC
Symbols for quantities are given in the International Standard ISO 31:1992, Quantities and
units.
In laboratory medicine, where lists of individual quantities are presented, each designation is
conventionally given in the exhaustive and unambiguous IUPAC/IFCC format System
Component; kind-of-quantity, where 'System' is the object under consideration, mostly having
a component of special interest, and 'kind-of-quantity' (or 'kind-of-property' if nominal
properties are included) is a label for general concepts such as length, diameter, and
amount-of-substance concentration. An example of an individual concept under quantity
could be 'Plasma (Blood) Sodium ion; amount-of-substance concentration equal to 143
mmol/l in a given person at a given time.
A vector or a tensor can also be a quantity if all its components are quantities.
1.2
(Note 2 to 1.1)
quantities of the same kind
quantities that can be placed in order of magnitude relative to one another
NOTES
1 Quantities of the same kind within a given system of quantities have the same dimension.
2
The subdivision of quantities into quantities of the same kind is to some extent arbitrary. For
example, moment of force and energy are, by convention, not regarded as being of the same
kind, although they have the same dimension, nor are heat capacity and entropy.
EXAMPLES
1.3
(1.2)
system of quantities
set of quantities together with a set of non-contradictory equations relating those quantities
NOTE
Ordinal quantities, such as Rockwell C hardness, are usually not considered to be part of a
system of quantities because they are related to other quantities through empirical relations
only.
1.4
International System of Quantities
ISQ
system of quantities, together with the equations relating the quantities, on which the SI is
based
NOTE
At present the ISQ is published in the International Standard ISO 31:1992, Quantities
and units.
1.5
(1.3)
base quantity
quantity, chosen by convention, used in a system of quantities to define other quantities
NOTES
1 There exist no equations relating the base quantities of a system of quantities.
2
EXAMPLE
The base quantities corresponding to the base units of the International System of Units
(SI) are given in the note to 1.16.
1.6
(1.4)
derived quantity
quantity, in a system of quantities, defined as a function of base quantities
EXAMPLE
In a system of quantities having the base quantities length and mass, mass density is a
derived quantity defined as the quotient of mass and volume (length to the third power).
Quantities having the same dimension are not necessarily quantities of the same kind.
In deriving the dimension of a quantity, no account is taken of any numerical factor, nor of its
scalar, vector or tensor character.
The dimension of a base quantity is generally referred to as base dimension, and similarly for
a derived dimension.
EXAMPLES
a) In the ISQ, where L, M and T denote the dimensions of the base quantities length, mass, and
time, the dimension of force is LMT 2.
b) In the same system of quantities ML3 is the dimension of mass concentration and also of
volumic mass (mass density).
1.8
(1.6)
quantity of dimension one
dimensionless quantity
quantity for which all the exponents of the factors corresponding to the base quantities in the
representation of its dimension are zero
NOTES
1
The values of quantities of dimension one are simply numbers.
2
The term dimensionless quantity is for historical reasons commonly used. It stems from the
fact that all exponents are zero in the symbolic representation of the dimension for such
quantities. However, the term quantity of dimension one reflects the convention in which the
symbolic representation of the dimension for such quantities is the symbol 1 (see ISO 310 :1992, subclause 2.2.6).
EXAMPLES
Plane angle, solid angle, linear strain, friction factor, refractive index, mass fraction,
amount-of-substance fraction, Mach number, Reynolds number, degeneracy in quantum
mechanics, number of turns in a coil, number of molecules.
1.9
(1.7)
unit
measurement unit
Units of quantities of the same dimension may be designated by the same name and
symbol even when the quantities are not of the same kind. For example the joule per
kelvin, J/K, is the name and symbol of both a unit of heat capacity and a unit of entropy,
which are generally not considered to be quantities of the same kind.
Units of quantities of dimension one are simply numbers. In some cases these
numbers are given special names, e.g. radian and steradian, or are expressed by
quotients such as millimole per mole.
1.10
(1.18)
quantity value
value of a quantity
value
magnitude of a quantity represented by a number and a reference
NOTES
1
EXAMPLES
a) Length of a given rod:
5.34 m
or
0.152 kg
5 qC
(7 + 3j) :
1.52
f)
or
534 cm
152 g
3 g/kg or 310 9
1.76 mmol/kg
1.11
(1.13)
base unit
base measurement unit
1.12
(1.14)
derived unit
unit for a derived quantity
EXAMPLES
In a system of quantities where length and time are base quantities, the metre per
second, symbol m/s, the kilometre per hour, km/h, and the knot, equal to one nautical mile
per hour, are derived units of velocity.
1.13
(1.10)
coherent derived unit
derived unit that, for a given system of quantities and for a chosen set of base units, is a
product of powers of base units with the proportionality factor one
NOTES
1 Coherence can be determined only with respect to a particular system of quantities and a
given set of base units.
Examples
If the metre, the second, and the mole are base units, the metre per second is the coherent
derived unit of velocity when velocity is defined by the quantity equation v = dr/dt, and the mole
per cubic metre is the coherent derived unit of amount-of-substance concentration when
amount-of-substance concentration is defined by the quantity equation c = n/V. The kilometre
per hour and the knot, given as examples of derived units in 1.12, are not coherent derived units
in such a system.
A derived unit can be coherent with respect to one system of quantities, but not to another.
The coherent derived unit for every quantity of dimension one in any system of quantities
is the number one, symbol 1.
1.14
(1.9)
system of units
conventionally selected set of base units and derived units, and also their multiples and
submultiples, together with a set of rules for their use
1.15
(1.11)
coherent system of units
system of units, based on a given system of quantities, in which the unit for each derived
quantity is a coherent derived unit
NOTE
A system of units can be coherent only with respect to the base units and the system of
equations that define the quantities involved.
EXAMPLES
a) The CGS system (based on the centimetre, gram, and second) in classical mechanics.
b) The International System of Units.
1.16
(1.12)
International System of Units
SI
coherent system of units based on the ISQ, their names and symbols, and a series of
prefixes and their names and symbols, together with rules for their use, adopted by the
General Conference on Weights and Measures (CGPM)
NOTE
The SI is founded on the following seven base quantities and base units:
_______________________________________________________
Base quantity
Base unit
Name
Name
Symbol
_______________________________________________________
length
metre
mass
kilogram
kg
time
second
electric current
ampere
thermodynamic temperature
kelvin
amount of substance
mole
mol
luminous intensity
candela
cd
_______________________________________________________
1.17
(1.16)
multiple of a unit
unit formed from a given unit by multiplying by an integer greater than one
EXAMPLES
a) The kilometre is a decimal multiple of the metre.
b) The hour is a non-decimal multiple of the second.
1.18
(1.17)
submultiple of a unit
unit formed from a given unit by dividing by an integer greater than one
EXAMPLES
a) The millimetre is a decimal submultiple of the metre.
b) The arc second is a non-decimal submultiple of the arc minute.
1.19
(1.20)
conventional quantity value
conventional value of a quantity
conventional value
value attributed by formal agreement to a quantity for a given purpose
EXAMPLES
a) Standard acceleration of free fall, gn = 9.806 65 m s2.
NOTE
Standard acceleration of free fall was formerly called standard acceleration due to gravity.
b) The conventional value for the Josephson constant, KJ-90 = 483 597.9 GHz V1.
1.20
(1.21)
numerical quantity value
numerical value of a quantity
numerical value
number in the representation of a quantity value
NOTE
For quantities that have a unit, the numerical value is frequently denoted { Q} =Q/[Q],
where Q is the symbol for the quantity and [Q] is the symbol for the unit.
1.21
quantity equation
equation relating quantities
EXAMPLES
a) Q = ] Q1 Q2, or {Q} [Q] = ] {Q1} [Q1] {Q2} [Q2], or {Q} [Q] = ] {Q1} {Q2} [Q1] [Q2], where {Q},
{Q1},and {Q2} denote the numerical values of the quantities Q, Q1, and Q2 expressed in terms
of units [Q], [Q1], and [Q2], respectively, and where ] is a numerical factor.
b) T = (1/2) mv2, where T is the kinetic energy, m is the mass, v the speed of a specified particle.
c) n = I.t / F where n is the amount of substance, I the electric current, t the duration, and F the
Faraday constant.
10
1.22
unit equation
equation relating units
EXAMPLES
a) [Q] = [Q1] [Q2] for the quantity equation given in the first example of 1.21 under the condition
that the units [Q], [Q1], and [Q2] are coherent.
b) J = kg m2 s2 , where J, kg, m, and s are the symbols for the joule, kilogram, metre, and
second, respectively.
c) [U%@ g/l, where [U%@is a symbol for a unit of mass concentration of substance B, and g and l
are the symbols for the gram and litre.
1.23
numerical value equation
numerical quantity value equation
equation relating numerical quantity values
EXAMPLES
a) {Q} = ] {Q1} {Q2} for the quantity equation given in the first example of 1.21 under the condition
that the units [Q], [Q1], and [Q2] are coherent.
b) In the equation for kinetic energy of a particle, T = (1/2) mv2 , if m = 2 kg and v = 3 m/s, so that
T = 9 J, then 9 = (1/2) u 2 u 32 is a numerical value equation.
1.24
quantity calculus
formalism for algebraic manipulation of symbols representing quantities
NOTE
In this formalism, equations between quantities have the advantage over equations between
numerical values of being independent of the choice of units (see ISO 31-0:1992, subclause
2.2.2).
1.25
conversion factor between units
ratio of two units for quantities of the same kind
EXAMPLES
a) km/m = 1000 and thus 1 km = 1000 m
b) h/s = 3600 and thus 1 h = 3600 s
c) (km/h)/(m/s) = (1/3.6) and thus 1 km/h = (1/3.6) m/s
1.26
ordinal quantity
11
Ordinal quantities can enter into empirical relations only and have no dimension.
Ordinal quantities are arranged according to conventional reference measurement
scales (see 2.34).
EXAMPLES
a) Rockwell C hardness
b) Octane number for petroleum fuel
c) Earthquake strength on the Richter scale
12
2.2
(2.2)
metrology
field of knowledge concerned with measurement
NOTE
Metrology includes all theoretical and practical aspects of measurement, whichever the
measurement uncertainty and field of application.
2.3
(2.6)
measurand
quantity intended to be measured
NOTES
1 The measurement can change the phenomenon, body, or substance under study such that
the quantity that is actually measured differs from the measurand.
Example
The potential difference between the terminals of a battery may decrease when
using a voltmeter with a significant internal conductance to perform the
measurement. The open-circuit potential difference can be calculated from the
internal resistances of the battery and the voltmeter.
2
Observe that this definition differs from that in VIM, 2nd Edition, Item 2.6, and some other
vocabularies, that define the measurand as the quantity subject to measurement.
The description of a measurand requires specification of the state of the phenomenon, body,
or substance under study.
2.4
(2.3)
measurement principle
principle of measurement
phenomenon serving as the basis of a measurement
13
NOTE
The measurement principle can be a physical, chemical, or biological phenomenon.
EXAMPLES
a) Thermoelectric effect applied to the measurement of temperature.
b) Energy absorption applied to the measurement of amount-of-substance concentration.
c) Lowering of the concentration of glucose in blood in a fasting rabbit applied to the measurement of
insulin concentration in a preparation.
2.5
(2.4)
measurement method
method of measurement
generic description of a logical sequence of operations used in a measurement
NOTES
1. Measurement methods may be qualified in various ways such as:
x substitution measurement method
x differential measurement method
x null measurement method
x direct measurement method
x indirect measurement method
See IEC 60050-300:2001
2. A measurement often requires the sequential or parallel use of several pieces of equipment or
reagents or both. Then the measurement method consists of a short presentation of the
procedural structure.
2.6
(2.5)
measurement procedure
detailed description of a measurement according to one or more measurement principles
and to a given measurement method
NOTE
The measurement procedure is usually documented in sufficient detail to enable an operator
to perform a measurement.
2.7
primary measurement procedure
primary procedure
measurement procedure used to realize the definition of a measurement unit and obtain
the quantity value and measurement uncertainty of a primary measurement standard
NOTES
1
CCQM uses the term direct primary method of measurement for this concept.
14
Definitions of two subordinate concepts, which could be termed direct primary reference
measurement procedure and ratio primary reference measurement procedure, are given
by CCQM (5th Meeting, 1999).
2.8
measurement scale
ordered set of values of quantities of a given kind, continuous or discrete, used in arranging
quantities of the same kind by magnitude
2.9
(1.22)
conventional reference measurement scale
conventional reference scale
measurement scale, defined by general agreement
NOTES
1
EXAMPLES
a) Rockwell C hardness scale;
b) pH scale in chemistry;
c) scale of octane numbers for petroleum fuel;
d) International Temperature Scale of 1990 (ITS-90).
2.10
(3.1)
measurement result
result of measurement
information about the magnitude of a quantity, obtained experimentally
NOTES
1 The information consists of a set of quantity values reasonably being attributed to the
measurand, usually summarized as a single quantity value and a measurement
uncertainty. The single quantity value is an estimate, often an average or the median of the
set.
2
If the measurand is considered to be sufficiently well described by a single quantity value (see
GUM, 1993, 1.2), it is common practice to have the term measurement result comprise the
estimated value only. The measurement uncertainty associated with this measurement result
is then stated separately.
15
The parameter may be, for example, a standard deviation called standard measurement
uncertainty (or a given multiple of it), or the half-width of an interval, having a stated
coverage probability.
It is understood that the quantity value of a measurement result is the best estimate of the
value of the measurand (see Note 2 of 2.10), and that all components of measurement
uncertainty, including those arising from systematic effects, such as components associated
with corrections and the assigned values of measurement standards, contribute to the
dispersion.
2.12
definitional measurement uncertainty
definitional uncertainty
component of measurement uncertainty resulting from the inherently finite amount of detail
in the definition of a measurand
NOTES
1
Any change in the descriptive detail of a measurand, through the corresponding change in
the measurement function, produces a new measurand having a new definitional
measurement uncertainty.
16
Defining the measurand is the first step of any measurement procedure. The ensuing
definitional uncertainty can therefore be considered as a part of the measurement
uncertainty.
The definitional uncertainty is a lower limit for the measurement uncertainty (see GUM).
2.13
Type A evaluation of measurement uncertainty
Type A evaluation
method of evaluation of a component of measurement uncertainty by a statistical analysis
of quantity values obtained by measurements under repeatability conditions
NOTE
For information about statistical analysis, see the GUM.
2.14
Type B evaluation of measurement uncertainty
Type B evaluation
method of evaluation of a component of measurement uncertainty by means other than a
statistical analysis of quantity values obtained by measurement
EXAMPLES
The component of measurement uncertainty may be:
x
x
x
x
x
2.15
standard measurement uncertainty
standard uncertainty of measurement
standard uncertainty
measurement uncertainty expressed as a standard deviation
NOTE
Sometimes the standard measurement uncertainty characterizing a measurement result is
obtained by taking into account standard measurement uncertainties and covariances of the
input quantities to the measurement function. This standard measurement uncertainty is
termed combined standard uncertainty in the GUM.
2.16
combined standard measurement uncertainty
combined standard uncertainty
standard measurement uncertainty of the measurement result when that result is
obtained from the values of a number of other quantities, equal to the positive square root
of a sum of terms, the terms being the variances and/or covariances of the values of these
17
2.17
coverage factor
number by which a standard measurement uncertainty of a measurement result is
multiplied to obtain an expanded measurement uncertainty
2.18
expanded measurement uncertainty
expanded uncertainty
half-width of a symmetric coverage interval, centered around the estimate of a quantity,
with a specified coverage probability
NOTES
1 Expanded measurement uncertainty is defined only for unimodal, symmetric probability
density functions.
2
2.19
coverage interval
interval of values which can be attributed to a quantity and, based on the available
information, is associated with a stated high probability
2.20
coverage probability
probability associated with a coverage interval
NOTE
The coverage probability is sometimes termed level of confidence (see GUM).
2.21
target measurement uncertainty
target uncertainty
measurement uncertainty formulated as a goal and decided on the basis of a specified
intended use of measurement results
2.22
(6.11)
calibration of a measuring system
18
Definition (b) is consistent with the definition of calibration in IEC 60050-300, item 311-01-09.
2.23
calibration hierarchy
sequence of calibrations of measuring systems between a stated metrological reference
and the final measuring system
NOTES
1
2.24
(6.10)
metrological traceability
property of a measurement result relating the result to a stated metrological reference
through an unbroken chain of calibrations of a measuring system or comparisons, each
contributing to the stated measurement uncertainty
NOTES
1 For this definition, a stated metrological reference can be a definition of a measurement unit
through its practical realization, or a measurement procedure, or a measurement standard.
2
19
Specification of the stated metrological reference must include the time at which the stated
metrological reference was used when establishing the calibration hierarchy.
The abbreviated term traceability is sometimes used for metrological traceability as well as
for other concepts, such as sample traceability or document traceability or instrument
traceability, where the history (trace) of an item is meant. Therefore, the full term should be
preferred.
For measurements with more than one input quantity to the measurement function, each
of the input quantities should itself be metrologically traceable.
2.25
metrological traceability chain
traceability chain
chain of alternating measuring systems with associated measurement procedures and
measurement standards, from a measurement result to a stated metrological reference
NOTE
A metrological traceability chain is defined through a calibration hierarchy from the
measurement result to the stated metrological reference.
2.26
metrological traceability to a measurement unit
metrological traceability to a unit
metrological traceability of a measurement result to the definition of a measurement unit
through a stated metrological traceability chain
NOTE
The phrase traceability to the SI is sometimes used; it means metrological traceability to a
unit of the International System of Units.
2.27
verification
confirmation through examination of a given item and provision of objective evidence that it
fulfils specified requirements
[modified from ISO 9000:2000, item 3.8.4]
NOTE
Verification should not be confused with calibration of a measuring system, or vice versa.
EXAMPLES
a) Demonstration that a given reference material as claimed is homogeneous down to samples
having a mass of 10 mg for the quantity and measurement procedure concerned.
b) Demonstration that stated performance properties of a measuring system are achieved.
2.28
validation
20
confirmation through examination of a given item and provision of objective evidence that it
fulfils the requirements for a stated intended use
[modified from ISO 9000:2000, item 3.8.5]
EXAMPLE
A measurement procedure, ordinarily used for the measurement of nitrogen concentration
in water, may be validated also for the measurement of nitrogen concentration in human
serum.
2.29
comparability of measurement results
comparability
property of measurement results enabling them to be compared because they are
metrologically traceable to the same stated metrological reference
NOTE
Comparability does not necessitate that the quantity values compared are of the same order
of magnitude.
EXAMPLES
a) Measurement results, for the distances from Earth to Moon and from Paris to London, are
comparable when (which is often the case) they are both metrologically traceable to the same
unit, for instance the metre.
b) Measurement results for amount-of-substance concentrations of a normal and a dangerous
level of cholesterol in human serum are comparable when they are both metrologically
traceable to the same unit, for instance mole per litre.
2.30
compatibility of measurement results
compatibility
property satisfied by all the measurement results of the same quantity, characterized by an
adequate overlap of their corresponding sets of quantity values
NOTE
See IEC 60050-300:2001, item 311-01-14.
2.31
(2.7)
influence quantity
quantity which, in a direct measurement, is neither the measurand nor the quantity being
measured, but whose change affects the relation between the indication of the measuring
system and the measurement result
NOTES
1
21
In the GUM, the term influence quantity covers both the sense given in the above
definition and a quantity that is not the measurand but that affects the measurement
result.
EXAMPLES
a) frequency in the measurement of the amplitude of an alternating current;
b) bilirubin amount-of-substance concentration in the measurement of haemoglobin concentration in
human blood plasma;
c) temperature of a micrometer used for measurement of length of a rod, but not the temperature of
the rod itself;
d) temperature in the measurement of pH;
e) background pressure in the ion source of a mass spectrometer during a measurement of amountof-substance fraction.
2.32
(3.14)
correction
modification applied to a quantity value obtained from measurement, to compensate for a
systematic effect
NOTES
1
2
2.33
measurement function
function expressing the mathematical relation between one or more measurands and the
quantity or quantities that must be measured, or whose values can be otherwise obtained,
to calculate a value of each measurand
NOTE
The general measurement function, f, for a single measurand, Y, can be expressed as Y =
f(X1, X2, , Xn), where X1, X2, , Xn are the input quantities to the measurement
function, as explained in the GUM (4.1).
2.34
input quantity to a measurement function
quantity that must be measured, or whose value can be otherwise obtained, to calculate a
value of a measurand as an output of the measurement function
22
2.35
measurement precision
precision
closeness of agreement between quantity values obtained by replicate measurements of a
quantity, under specified conditions
NOTE
Measurement precision is usually expressed numerically by measures of imprecision, such as
standard deviation, variance, or coefficient of variation under the specified conditions of
measurement.
2.36
(3.6 NOTE 2)
repeatability condition of measurement
repeatability condition
condition of measurement in a set of conditions including the same measurement
procedure, same operator, same measuring system, same operating conditions and same
location, and replicated measurements over a short period of time
2.37
(3.6)
measurement repeatability
repeatability
measurement precision under repeatability conditions of measurement
2.38
intermediate precision condition of measurement
intermediate precision condition
condition of measurement in a set of conditions including the same measurement
procedure, same location, and replicated measurements over an extended period of time
NOTES
1 Changes within a given type of calibration, operator, and measuring system are allowed.
2
A specification should contain the conditions changed and unchanged, to the extent practical.
2.39
23
2.41
(3.7)
measurement reproducibility
reproducibility
measurement precision under reproducibility conditions of measurement
24
NOTE
Selectivity used in physics is a concept close to specificity used in chemistry.
25
3.1
(4.1)
measuring instrument
device or combination of devices designed for measurement of quantities
3.2
(4.2)
material measure
device reproducing or supplying, in a permanent manner during its use, quantities of given
kinds, each with an assigned value
EXAMPLES
a) A weight.
b) A volume measure (supplying one or several quantity values, with or without a measurement
scale).
c) A standard electric resistor.
d) A line scale (ruler).
e) A gauge block.
f)
g) A reference material.
3.3
(4.3)
measuring transducer
device that provides at its output a quantity having a determined relation to the quantity at its
input
EXAMPLES
a)
Thermocouple.
b)
Current transformer.
c)
Strain gauge.
d)
pH electrode.
e)
Bourdon tube.
f)
Bimetal strip.
26
3.4
(4.4)
measuring chain
series of elements of a measuring system constituting one single path of the measurement
signal
EXAMPLE
An electro-acoustic measuring chain comprising a microphone, attenuator, filter, amplifier, and
voltmeter.
3.5
(4.5)
measuring system
measurement system
set of measuring instruments and other devices or substances assembled and adapted to
the measurement of quantities of specified kinds within specified intervals of values
NOTE
The substance can be a chemical reagent.
3.6
(4.6)
indicating measuring instrument
measuring instrument providing an output signal carrying information about the value of
the quantity to be measured
NOTES
1
3.7
(4.12)
displaying device
device providing the indication of a measuring system in visual form
NOTE
A displaying device may be the device by which the quantity value supplied by a material
measure is displayed or set.
3.8
(4.14)
sensor
element of a measuring system that is directly affected by the phenomenon, body, or
substance carrying the quantity to be measured
27
NOTE
In some fields the term detector is used for this concept.
EXAMPLES
a) Measuring junction of a thermoelectric thermometer.
b) Rotor of a turbine flow meter.
c) Bourdon tube of a pressure gauge.
d) Float of a level-measuring instrument.
e) Photocell of a spectrophotometer.
f)
3.9
(4.15)
detector
device or substance that indicates the presence of a phenomenon, body, or substance when
a threshold value of an associated quantity is exceeded
NOTES
1
2
In some fields the term detector is used for the concept of sensor.
In chemistry the term indicator is frequently used for this concept.
EXAMPLES
a) Halogen leak detector.
b) Litmus paper.
3.10
(4.17)
scale of a displaying device
scale
part of a displaying device consisting of an ordered set of marks, together with any
associated numbers or quantity values
NOTE
Related terms and definitions can be found in IEC 60050-300, Part 314.
3.11
(4.30)
adjustment of a measuring system
adjustment
set of operations carried out on a measuring system in order that it provide prescribed
indications corresponding to given values of the quantity to be measured
NOTE
28
3.12
zero adjustment of a measuring system
zero adjustment
adjustment of a measuring system providing a null indication corresponding to a null
value of the quantity to be measured
29
The indication and the value of the quantity being measured are not necessarily values of
quantities of the same kind.
4.2
(4.19)
indication interval
set of quantity values bounded by the extreme possible indications of a measuring
system
NOTES
1 An indication interval is usually stated in terms of its smallest and greatest quantity values, for
example, 99 V to 201 V.
2
4.3
(5.1)
nominal indication interval
nominal interval
set of quantity values, bounded by rounded or approximate extreme indications obtainable
with a particular setting of the controls of a measuring system and used to designate this
setting
NOTES
1 A nominal interval is usually stated as its smallest and greatest quantity values, for example, 100
V to 200 V". Where one of the extreme quantity values is zero, the nominal interval is sometimes
stated solely as the greatest quantity value, for example a nominal interval of 0 V to 100 V is
expressed as 100 V.
2
4.4
(5.2)
span of a nominal interval
span
30
4.5
(5.3)
nominal quantity value
nominal value
rounded or approximate quantity value of a characteristic of a measuring system that
provides a guide to its use
EXAMPLES
a) 100 : as the quantity value marked on a standard resistor.
b) 100 cl as the value marked on a single-mark volumetric flask.
c) 0.1 mol/l as the value for amount-of-substance concentration of a solution of hydrogen
chloride, HCl.
d) 25 C as the set point of a thermostatically controlled bath.
e) 40 C as a maximum temperature for storage.
4.6
(5.4)
measuring interval
working interval
set of values of the quantities of the same kind that can be measured by a given
measuring system, with specified measurement uncertainty under defined conditions
NOTE
In some fields the term is measuring range
4.7
steady state condition for a measuring system
steady state condition
operating condition of a measuring system in which the possible variation with time of the
quantity being measured is such that a calibration of the measuring system carried out
with a measurand constant with time remains valid
4.8
(5.5)
rated operating condition for a measuring system
rated operating condition
condition that must be fulfilled during measurement in order that a measuring system
perform as designed
31
4.9
(5.6)
limiting condition for a measuring system
limiting condition
extreme condition that a measuring system is required to withstand without damage, and
without degradation of specified metrological characteristics when it is subsequently
operated under its rated operating conditions
NOTES
1
2
4.10
(5.7 )
reference condition for a measuring system
reference condition
condition of use prescribed for evaluating the performance of a measuring system or for
comparison of measurement results
NOTES
1 Reference conditions generally specify intervals of values for any influence quantity.
2
In IEC 60050-300, item 311-06-02, this term is used for a concept related to optimum
condition.
4.11
optimum condition for a measuring system
optimum condition
condition of use for a measuring system, under which its contribution to measurement
uncertainty is minimum
NOTE
In IEC 60050-300, item 311-06-02, a related concept is termed reference conditions.
4.12
(5.10)
sensitivity of a measuring system
sensitivity
quotient of the change in the indication of a measuring system and the corresponding
change in the value of the quantity being measured
NOTES
1 The sensitivity can depend on the value of the quantity being measured.
32
The change considered in the value of the quantity being measured must be large compared with
the resolution of the measuring system.
4.13
resolution of a measuring system
smallest change, in the value of a quantity being measured by a measuring system, that
causes a perceptible change in the corresponding indication
NOTE
The resolution of a measuring system may depend on, for example, noise (internal or
external) or friction. It may also depend on the value of the quantity being measured.
4.14
(5.12)
resolution of a displaying device
smallest difference between indications of a displaying device that can be meaningfully
distinguished
4.15
(5.11)
discrimination threshold
largest change in the value of a quantity being measured by a measuring system that
causes no detectable change in the corresponding indication
NOTE
The discrimination threshold may depend on, for example, noise (internal or external) or
friction. It may also depend on the value of the quantity being measured.
4.16
(5.13)
dead band of a measuring system
dead band
maximum interval through which the value of a quantity being measured by a measuring
system can be changed in both directions without producing a detectable change in the
corresponding indication
NOTE
The dead band can depend on the rate of change.
4.17
(5.14)
stability of a measuring system
stability
ability of a measuring system to maintain its metrological characteristics constant with time
33
4.19
variation due to an influence quantity
difference between the indications of a measuring system for the same value of the
quantity being measured when an influence quantity assumes, successively, two different
values
NOTE
For a material measure, the variation due to an influence quantity is the difference between
the values of the supplied quantity when the influence quantity assumes two different values.
4.20
(5.17)
step-change response time of a measuring system
step-change response time
duration between the instant when a quantity value at the input of a measuring system is
subjected to a step change between two specified steady states and the instant when the
corresponding indication settles within specified limits around its final steady value
4.21
(5.27)
repeatability of a measuring system
repeatability
property of a measuring system to provide closely similar indications for replicated
measurements of the same quantity under repeatability conditions
NOTE
Repeatability can be expressed quantitatively in terms of the dispersion parameters of the
indications of the measuring system.
4.22
instrumental uncertainty
component of measurement uncertainty attributed to a measuring instrument and
determined by its calibration
NOTES
34
This definition is consistent with the IEC terminology, for instance IEC 60050-300, item 31103-09.
Instrumental uncertainty is useful when the measuring instrument is used to perform a single
measurement, however, caution must be taken not to overestimate or underestimate the
measurement uncertainty when repeated measurements are performed.
4.23
accuracy class
<uncertainty approach> class of measuring instruments that meet stated metrological
requirements which are intended to keep instrumental uncertainty within specified limits
under specified operating conditions
NOTE
Related concepts are accuracy class <classical approach> and accuracy of measurement.
35
In science and technology, the English word standard is used with two different
meanings: as a widely adopted written standard, specification, technical
recommendation or similar document (in French norme) and as a measurement
standard (in French talon). This Vocabulary is concerned solely with the second
meaning.
5.1
(6.1)
measurement standard
standard
etalon
realization of the definition of a given quantity, with stated value and measurement
uncertainty, used as a reference
NOTES
1 The realization of the definition of a given quantity can consist of a measuring system, a
material measure, or a reference material.
2
Several quantities of the same kind or of different kinds may be realized in one device which
is commonly also called a measurement standard.
The word embodiment is sometimes used in the English language instead of realization.
EXAMPLES
a) 1 kg mass standard.
b) 100 standard resistor.
c) Standard ammeter.
d) Caesium frequency standard.
e) Standard hydrogen electrode.
f)
g) Reference material providing certified values with measurement uncertainties for the mass
concentration of each of ten different proteins.
36
5.2
(6.3)
national measurement standard
national standard
measurement standard designated as a national stated metrological reference
5.3
(6.4)
primary measurement standard
primary standard
measurement standard whose quantity value and measurement uncertainty are
established without relation to another measurement standard for a quantity of the same
kind
NOTES
1
This definition implies that the establishment of a primary measurement standard has to
refer to the definition of any unit concerned, particularly an SI unit.
EXAMPLES
a) Primary measurement standard of amount-of-substance concentration prepared by
dissolving a known amount of substance of a chemical component to a known volume
of solution.
b) Primary measurement standard for pressure based on separate measurements of
force and area.
c) Primary measurement standard for isotope amount-of-substance ratio
measurements, prepared by mixing known amounts of substance of specified
isotopes.
5.4
(6.5)
secondary measurement standard
secondary standard
measurement standard whose quantity value and measurement uncertainty are
assigned through calibration against, or comparison with, a primary measurement
standard for a quantity of the same kind
NOTES
1
The relation may be obtained directly between the primary measurement standard and
the secondary measurement standard, or involve an intermediate measuring system
calibrated by the primary standard and assigning a measurement result to the secondary
standard.
A measurement standard having its quantity value assigned by a ratio primary
measurement procedure is a secondary measurement standard.
37
5.5
(6.6)
reference measurement standard
reference standard
measurement standard used for the calibration of working measurement standards in a
given organization or at a given location
5.6
(6.7)
working measurement standard
working standard
measurement standard that is used routinely to calibrate, verify, or check measuring
systems, material measures, or reference materials
NOTE
A working measurement standard is usually calibrated with a reference measurement standard.
5.7
(6.9)
travelling measurement standard
travelling standard
measurement standard, sometimes of special construction, intended for transport between
different locations
EXAMPLE
A portable battery-operated caesium 133 frequency standard
5.8
(6.8)
transfer measurement device
transfer device
measurement device used as an intermediary to compare measurement standards
5.9
intrinsic measurement standard
intrinsic standard
measurement standard based on a sufficiently stable and reproducible property of a
phenomenon or substance
NOTES
1
The quantity value of an intrinsic standard is assigned by consensus and does not need to
be established by relating it to another measurement standard of the same type. Its
measurement uncertainty is determined by considering two components: (A) that
associated with its consensus quantity value and (B) that associated with its construction,
implementation and maintenance.
38
The quantity value of an intrinsic standard may depend on the consensus value of a
fundamental constant.
The stability and measurement reproducibility of some intrinsic standards come from the
fact that the phenomenon used is a quantum phenomenon.
The adjective intrinsic does not mean that this standard may be used without special care or
that this standard is immune to spurious effects.
EXAMPLES
a)
Triple-point-of-water cell as an intrinsic standard of thermodynamic temperature.
b)
c)
5.10
(6.7 NOTE)
check measurement device
check device
measurement device used routinely for verifying the functioning of a measuring system
5.11
(6.12)
conservation of a measurement standard
operation or set of operations necessary to preserve the metrological properties of a
measurement standard within stated limits
NOTE
Conservation commonly includes periodic verification or calibration, storage under
suitable conditions, and specified care in use.
5.12
calibrator
measurement standard used in the calibration of a measuring system
5.13
(6.13)
reference material
RM
material, sufficiently homogeneous and stable with respect to one or more specified
quantities, used for the calibration of a measuring system, or for the assessment of a
measurement procedure, or for assigning values and measurement uncertainties to
quantities of the same kind for other materials
39
NOTES
1 The term reference material designates a family of materials without necessarily implying a
hierarchy according to the magnitude of the measurement uncertainty.
2
Reference material comprises both precision control material, which needs not have an
assigned quantity value, and measurement standard functioning as trueness control
material or calibrator.
A reference material can be in the form of, e.g. a pure or mixed gas, liquid, solid or
suspension.
The term reference material is also used for materials realizing nominal properties, such as
colour.
EXAMPLES
a)
Water for the calibration of viscometers.
b)
c)
Human serum containing cholesterol without assigned quantity value, used only as a
precision control material.
5.14
(6.14)
certified reference material
CRM
reference material, accompanied by an authenticated certificate, having for each specified
quantity a value, measurement uncertainty, and stated metrological traceability chain
NOTES
1
Certified reference materials are generally prepared in batches. For a given batch, quantity
values and measurement uncertainties are obtained by measurements on samples
representative of the batch.
The quantity values assigned to a certified reference material are sometimes conveniently and
reliably obtained when the material is incorporated into a specially fabricated device. The
quantity value is sometimes the output of a device. Such devices may also be considered as
CRMs.
EXAMPLES
a)
b)
c)
d)
Procedures for the production and certification of certified reference materials are given in ISO
Guides 34 and 35.
40
Some reference materials and certified reference materials have quantities which, because
they cannot be correlated with an established chemical structure or for other reasons, cannot
be measured according to measurement procedures giving measurement results that are
metrologically traceable to measurement units of the International System of Units (SI) or
another system of units. Such materials include certain biological materials such as
vaccines to which International Units (IU) have been assigned by the World Health
Organization.
5.15
commutability of a reference material
property of a given reference material demonstrated by the closeness of agreement
between the relation among the measurement results, for a stated quantity in this material,
obtained according to two given measurement procedures, and the relation obtained
among the measurement results for other specified materials
NOTES
1
2
5.16
reference data
data that is critically evaluated and verified, obtained from an identified source, and related to
a property of a phenomenon, body, or substance, or a system of components of known
composition or structure
NOTE
Here the word data is used in the singular sense, instead of using the word datum.
5.17
standard reference data
reference data issued by a stated recognized authority
EXAMPLE
CODATA (1998) recommended values of the fundamental physical constants.
5.18
reference quantity value
reference value
quantity value, generally accepted as having a suitably small measurement uncertainty, to
be used as a basis for comparison with values of quantities of the same kind
NOTES
1 A reference quantity value with associated measurement uncertainty is usually referred to
x a material, e.g. a certified reference material,
x a device, e.g. a stabilized laser,
x a reference measurement procedure.
41
42
A1
(1.19)
true value of a quantity
true value
quantity value consistent with the definition of a quantity
NOTES
1
Within the Classical Approach a unique quantity value is thought to describe the
measurand. The true value would be obtainable by a perfect measurement, that is a
measurement without measurement error. The true value is by nature unobtainable.
Due to definitional measurement uncertainty, there is a distribution of true values
consistent with the definition of a measurand. This distribution is by nature unknowable.
The concept of true value is avoided in the Uncertainty Approach (D.3.5 in the GUM).
A2
(3.5)
accuracy of measurement
accuracy
closeness of agreement between a quantity value obtained by measurement and the true
value of the measurand
NOTES
1
The term accuracy of measurement should not be used for trueness of measurement
and the term measurement precision should not be used for accuracy of measurement.
A3
(5.18)
accuracy of a measuring system
accuracy
ability of a measuring system to provide a quantity value close to the true value of a
measurand
NOTES
1
Accuracy is greater when the quantity value is closer to the true value.
A4
trueness of measurement
43
The term trueness of measurement should not be used for accuracy of measurement.
A5
(3.10)
error of measurement
error
difference of quantity value obtained by measurement and true value of the measurand
NOTE
It is necessary to distinguish error of measurement from relative error of measurement.
A6
(5.20)
error of indication
difference of indication of a measuring system and true value of the measurand
NOTE
In the Classical Approach, indication of a measuring system and the value of the measurand
are assumed to be values of quantities of the same kind.
A7
(3.13)
random error of measurement
random error
difference of quantity value obtained by measurement and average that would ensue from
an infinite number of replicated measurements of the same measurand carried out under
repeatability conditions
NOTES
1
A.8
(3.14)
systematic error of measurement
44
Systematic error, and its causes, can be known or unknown. Correction should be applied
for systematic error, as far as it is known.
Systematic error equals the difference of error of measurement and random error of
measurement.
A9
(5.19)
accuracy class
<classical approach> class of measuring instruments that meet stated metrological
requirements which are intended to keep errors of indication within specified limits under
specified operating conditions
NOTES
1
2
A10
(5.21)
maximum permissible error
limit of error
one of the two extreme values of the error of indication permitted by specifications or
regulations for a given measuring system
A11
(5.22)
datum error of a measuring system
datum error
error of indication of a measuring system at a specified indication or a specified value of
the measurand
A12
(5.23)
zero error of a measuring system
zero error
datum error for zero value of the measurand
A13
(5.24)
45
46
!NNEXE "