90012
90012
90012
A07943 tneligA
1997-2012
Warranty
The material contained in this document is provided as is, and is subject to being changed, without notice, in future editions. Further, to the maximum extent permitted by applicable law, Agilent disclaims all warranties, either express or implied with regard to this manual and any information contained herein, including but not limited to the implied warranties of merchantability and fitness for a particular purpose. Agilent shall not be liable for errors or for incidental or consequential damages in connection with the furnishing, use, or performance of this document or any information contained herein. Should Agilent and the user have a separate written agreement with warranty terms covering the material in this document that conflict with these terms, the warranty terms in the separate agreement will control.
Safety Notices
Do not defeat power cord safety ground feature. Plug in to a grounded outlet. Do not use product in any manner not specified by the manufacturer. Do not install substitute parts or perform any unauthorized modification to the product. Return the product to an Agilent Technologies Sales and Service Office for service and repair to ensure that safety features are maintained.
No part of this manual may be reproduced in any form or by any means (including electronic storage and retrieval or translation into a foreign language) without prior agreement and written consent from Agilent Technologies as governed by the United States and international copyright laws.
WARNING A WARNING notice denotes a hazard. It calls attention to an operating procedure, practice, or the like that, if not correctly performed or adhered to, could result in personal injury or death. Do not proceed beyond a WARNING notice until the indicated conditions are fully understood and met.
Edition
Edition 5, May 2012
Printed in Malaysia Agilent Technologies, Inc. 815 14th Street S.W. Loveland, Colorado 80537 U.S.A.
Technologies Licenses
The hardware and/or software described in this document are furnished under a license and may be used or copied only in accordance with the terms of such license.
CAUTION A CAUTION notice denotes a hazard. It calls attention to an operating procedure, practice, or the like that, if not correctly performed or adhered to, could result in damage to the product or loss of important data. Do not proceed beyond a CAUTION notice until the indicated conditions are fully understood and met.
Assistance
Product maintenance agreements and other customer assistance agreements are available for Agilent Technologies products. For assistance, contact your nearest Agilent Technologies Sales and Service Office. Further information is available on the Agilent web site at www.agilent.com/find/assist.
Trademark Information
Microsoft and Windows are U.S. registered trademarks of Microsoft Corporation. All other brand and product names are trademarks or registered trademarks of their respective companies.
Symbols
Earth ground
Chassis ground
Certification
Agilent Technologies certifies that this product met its published specifications at the time of shipment from the factory. Agilent Technologies further certifies that its calibration measurements are traceable to the United States National Institute of Standards and Technology, to the extent allowed by that organization's calibration facility, and to the calibration facilities of other International Standards Organization members.
W ARN IN G Main power disconnect: Unplug product from wall outlet and remove power cord before servicing. Only qualified, service-trained personnel should remove the cover from the instrument or connect external wiring to a module. For continued protection against fire, replace the line fuse only with a fuse of the specified type and rating.
Note: Unless otherwise indicated, this manual applies to all serial numbers.
The Agilent Technologies 34970A combines precision measurement capability with flexible signal connections for your production and development test systems. Three module slots are built into the rear of the instrument to accept any combination of data acquisition or switching modules. The combination of data logging and data acquisition features makes this instrument a versatile solution for your testing requirements now and in the future. Convenient Data Logging Features Direct measurement of thermocouples, RTDs, thermistors, dc voltage, ac voltage, resistance, dc current, ac current, frequency, and period Interval scanning with storage of up to 50,000 time-stamped readings Independent channel configuration with function, Mx+B scaling, and alarm limits available on a per-channel basis Intuitive user interface with knob for quick channel selection, menu navigation, and data entry from the front panel Portable, ruggedized case with non-skid feet BenchLink Data Logger 3 Software for Microsoft Windows included Flexible Data Acquisition / Switching Features 612-digit multimeter accuracy, stability, and noise rejection Up to 60 channels per instrument (120 single-ended channels) Reading rates up to 600 readings per second on a single channel and scan rates up to 250 channels per second Choice of multiplexing, matrix, general-purpose Form C switching, RF switching, digital I/O, totalize, and 16-bit analog output functions GPIB (IEEE-488) interface and RS-232 interface are standard SCPI (Standard Commands for Programmable Instruments) compatibility Warning
The procedures in this manual are intended for use by qualified, service-trained personnel only.
Denotes a menu key. See the next page for details on menu operation.
8 9 10 11 12 13 14
Advanced Measurement / Utility Menus Low-Level Module Control Keys Single-Channel Monitor On/Off Key View Scanned Data, Alarms, Errors Menu Shift / Local Key Knob Navigation Arrow Keys
The menu will automatically timeout after about 20 seconds of inactivity. You will be returned to the operation in progress prior to entering the menu. 2 Press the same menu key again to move to the next item of the menu. Typically, this is where you choose parameter values for the selected operation.
3 Rotate the knob to view the choices on this level of the menu. When you reach the end of the list, rotate the knob in the opposite direction to view all of the other choices. The current selection is highlighted for emphasis. All other choices are dimmed.
4 Press the same menu key again to accept the change and exit the menu. A brief confirmation message is displayed.
Tip: To review the current configuration of a specific menu, press the menu key several times. A message NO CHANGES is displayed when you exit the menu. 3
Display Annunciators
SCAN MON VIEW CONFIG ADRS RMT ERROR EXT ONCE MEM LAST MIN MAX SHIFT 4W OC
Scan is in progress or enabled. Press and hold again to turn off. Monitor mode is enabled. Press again to turn off. Scanned readings, alarms, errors, or relay cycles are being viewed. Channel configuration is in progress on displayed channel. Measurement is in progress. Instrument is addressed to listen or talk over the remote interface. Instrument is in remote mode (remote interface). Hardware or remote interface errors are detected. Press to read errors. Instrument is configured for an external scan interval. Scan Once mode is enabled. Press to initiate and hold key to disable. Reading memory overflow; new readings will overwrite the oldest readings. Viewed data is the last reading stored during most recent scan. Viewed data is the minimum reading stored during most recent scan. Viewed data is the maximum reading stored during most recent scan. has been pressed. Press again to turn off. 4-wire function is in use on displayed channel. Offset compensation is enabled on displayed channel. Alarms are enabled on displayed channel. Mx+B scaling is enabled on displayed channel. HI or LO alarm condition has occurred on indicated alarms.
To review the display annunciators, hold down the turn on the instrument.
key as you
1 Slot Identifier (100, 200, 300) 2 Ext Trig Input / Alarm Outputs / Channel Advance Input / Channel Closed Output (for pinouts, see chapter 4 in Users Guide) 3 RS-232 Interface Connector
4 5 6 7
Power-Line Fuse-Holder Assembly Power-Line Voltage Setting Chassis Ground GPIB (IEEE-488) Interface Connector
Use the Menu to: Select the GPIB or RS-232 interface (see chapter 2 in Users Guide). Set the GPIB address (see chapter 2 in Users Guide). Set the RS-232 baud rate, parity, and flow control mode (see chapter 2 in Users Guide).
WARNING
For protection from electrical shock, the power cord ground must not be defeated. If only a two-contact electrical outlet is available, connect the instruments chassis ground screw (see above) to a good earth ground.
In This Book
Specifications Chapter 1 lists the technical specifications for the mainframe and plug-in modules. Quick Start Chapter 2 helps you get familiar with a few of the instruments front-panel features. Front-Panel Overview Chapter 3 introduces you to the front-panel menus and describes some of the instruments menu features. Calibration Procedures Chapter 4 provides calibration, verification, and adjustment procedures for the instrument. Theory of Operation Chapter 5 describes block and circuit level theory related to the operation the instrument. Service Chapter 6 provides guidelines for returning your instrument to Agilent Technologies for servicing, or for servicing it yourself. Replaceable Parts Chapter 7 contains detailed parts lists for the mainframe and plug-in modules. Schematics Chapter 8 contains the instruments block diagram, schematics, disassembly drawings, and component locator drawings.
If you have questions relating to the operation of the 34970A, call 1-800-452-4844 in the United States, or contact your nearest Agilent Technologies Sales Office. If your 34970A fails within one year of original purchase, Agilent will replace it free of charge. Call 1-800-829-4444 and select "Option 3" followed by "Option 1".
Contents
Chapter 1 Specifications DC, Resistance, and Temperature Accuracy Specifications 16 DC Measurement and Operating Characteristics 17 AC Accuracy Specifications 18 AC Measurement and Operating Characteristics 19 Measurement Rates and System Characteristics 20 Module Specifications 21 BenchLink Data Logger Software Specifications 24 Product and Module Dimensions 25 To Calculate Total Measurement Error 26 Interpreting Internal DMM Specifications 28 Configuring for Highest Accuracy Measurements 31 Chapter 2 Quick Start To Prepare the Instrument for Use 35 To Connect Wiring to a Module 36 To Set the Time and Date 38 To Configure a Measurement Channel 39 To Monitor a Single Channel 40 To Close a Channel 41 If the Instrument Does Not Turn On 42 To Adjust the Carrying Handle 44 To Rack Mount the Instrument 45 Chapter 3 Front-Panel Overview Front-Panel Menu Reference 49 To Unsecure for Calibration 51 To Secure Against Calibration 51 To Change the Security Code 52 Error Messages 52 To Perform a Zero Adjustment 53 To Apply Mx+B Scaling to Measurements 54 To Read the Relay Cycle Count 55 To Read a Digital Input Port 56 To Write to a Digital Output Port 57 To Read the Totalizer Count 58 To Output a DC Voltage 59
Contents
10
Contents
Chapter 4 Calibration Procedures Agilent Technologies Calibration Services 63 Calibration Interval 63 Adjustment is Recommended 63 Time Required for Calibration 64 Automating Calibration Procedures 64 Recommended Test Equipment 65 Input Connections 66 Calibration Security 67 To Unsecure the Instrument Without the Security Code 68 Calibration Message 69 Calibration Count 69 Calibration Procedure 70 Aborting a Calibration in Progress 70 Test Considerations 71 Performance Verification Tests 72 Self-Test 73 Quick Performance Check 74 Performance Verification Tests 74 Internal DMM Verification Tests 75 Zero Offset Verification 75 Gain Verification 77 Optional AC Performance Verification Tests 80 Internal DMM Adjustments 81 Zero Adjustment 81 Gain Adjustment 82 10 Vdc Adjustment Procedure (Optional) 85 Plug-in Module Test Considerations 87 Relay Verification 88 Relay Cycle Count 88 34901A Relay Contact Resistance Verification (Optional) 89 34902A Relay Contact Resistance Verification (Optional) 96 34903A Relay Contact Resistance Verification (Optional) 101 34904A Relay Contact Resistance Verification (Optional) 102 34905A/06A Relay Contact Resistance Verification (Optional) 105 34908A Relay Contact Resistance Verification (Optional) 106 Thermocouple Reference Junction (Optional) 112 Thermocouple Reference Junction Verification 112 Thermocouple Reference Junction Adjustments 113 34907A Analog Output 114 Analog Output Verification Test 114 Analog Output Adjustment 115
Contents
11
Contents
Chapter 5 Theory of Operation System Block Diagram 119 Floating Logic 120 Memory 123 Earth-Referenced Logic 124 Power Supplies 125 Front Panel 127 Backplane 128 Analog Bus 128 Digital Bus 128 Internal DMM 129 DMM Block Diagram 129 Input 130 Input Amplifier 131 Ohms Current Source 133 AC Circuit 134 A-to-D Converter 136 Switch Modules 138 Switch Module Control 138 Relay Drivers 140 34901A 142 34902A 144 34903A 146 34904A 147 34905A/34906A 148 34908A 149 Multifunction Module 151 Multifunction Control 151 Totalizer 153 Analog Output 154 Digital I/O 155 Chapter 6 Service Operating Checklist 159 Is the instrument inoperative? 159 Does the instrument fail self-test? 159 Is the Current measurement function inoperative? 159 Types of Service Available 160 Standard Repair Service (worldwide) 160 Express Exchange (U.S.A. only) 160 Repackaging for Shipment 161 Cleaning 161 Electrostatic Discharge (ESD) Precautions 162 Surface Mount Repair 162 To Replace the Power-Line Fuse 163
Contents
12
Contents
Chapter 6 Service (continued) Troubleshooting Hints 163 Unit is Inoperative 163 Unit Reports Error 705 164 Isolating to an Assembly 164 Unit Fails Self-Test 164 Power Supplies 165 Self-Test Procedures 167 Power-On Self-Test 167 Complete Self-Test 167 Plug-in Module Self-Test 167 Self-Tests 168 Battery Check and Replacement 172 To Verify the Battery 173 To Replace the Battery 173 Disassembly 174 General Disassembly 175 Internal DMM Disassembly 176 Front-Panel Chassis Disassembly 177 Additional Disassembly 178 Plug-in Module Disassembly 179 Chapter 7 Replaceable Parts Replaceable Parts 182 To Order Replaceable Parts 182 34970A Mainframe 183 34970-66501 Main PC Assembly (A1) 184 34970-66502 Front-Panel and Keyboard PC Assembly (A2) 189 34970-66503 Backplane PC Assembly (A3) 190 34970-66504 Internal DMM PC Assembly (A4) 191 34901A 20-Channel Multiplexer 196 34902A 16-Channel Multiplexer 200 34903A 20-Channel Actuator 202 34904A 4x8 Matrix 204 34905A/34906A RF Multiplexer 207 34907A Multifunction Module 209 34908A 40-Channel Multiplexer 213 Manufacturers List 216
Contents
13
Contents
Chapter 8 Schematics Agilent 34970A System Block Diagram 221 A1 Component Locator (top) 222 A1 Component Locator (bottom) 223 A1 Power Supply Schematic (Sheet 1 of 4) 224 A1 Floating Logic Schematic (Sheet 2 of 4) 225 A1 Earth Referenced Logic Schematic (Sheet 3 of 4) 226 A1 Memory Schematic (Sheet 4 of 4) 227 A2 Component Locator 228 A2 Display and Keyboard Schematic 229 A3 Component Locator 230 A3 Backplane Schematic 231 A4 Component Locator (top) 232 A4 Component Locator (bottom) 233 A4 Input and Protection Schematic (Sheet 1 of 4) 234 A4 Input Amplifier and Ohms Current Schematic (Sheet 2 of 4) 235 A4 AC Schematic (Sheet 3 of 4) 236 A4 A/D Converter Schematic (Sheet 4 of 4) 237 34901A 20-Channel Multiplexer Component Locator 238 34901A 20-Channel Multiplexer Schematic (Sheet 1 of 5) 239 34901A 20-Channel Multiplexer Schematic (Sheet 2 of 5) 240 34901A 20-Channel Multiplexer Schematic (Sheet 3 of 5) 241 34901A 20-Channel Multiplexer Schematic (Sheet 4 of 5) 242 34901A 20-Channel Multiplexer Schematic (Sheet 5 of 5) 243 34902A 16-Channel Multiplexer Component Locator 244 34902A 16-Channel Multiplexer Schematic (Sheet 1 of 4) 245 34902A 16-Channel Multiplexer Schematic (Sheet 2 of 4) 246 34902A 16-Channel Multiplexer Schematic (Sheet 3 of 4) 247 34902A 16-Channel Multiplexer Schematic (Sheet 4 of 4) 248 34903A 20-Channel Actuator Component Locator 249 34903A 20-Channel Actuator Schematic (Sheet 1 of 3) 250 34903A 20-Channel Actuator Schematic (Sheet 2 of 3) 251 34903A 20-Channel Actuator Schematic (Sheet 3 of 3) 252 34904A 4x8 Matrix Component Locator 253 34904A 4x8 Matrix Schematic (Sheet 1 of 3) 254 34904A 4x8 Matrix Schematic (Sheet 2 of 3) 255 34904A 4x8 Matrix Schematic (Sheet 3 of 3) 256 34905A/34906A RF Multiplexer Component Locator 257 34905A/34906A RF Multiplexer Schematic (Sheet 1 of 2) 258 34905A/34906A RF Multiplexer Schematic (Sheet 2 of 2) 259 34907A Multifunction Module Component Locator 260 34907A Multifunction Module Schematic (Sheet 1 of 5) 261 34907A Multifunction Module Schematic (Sheet 2 of 5) 262 34907A Multifunction Module Schematic (Sheet 3 of 5) 263 34907A Multifunction Module Schematic (Sheet 4 of 5) 264 34907A Multifunction Module Schematic (Sheet 5 of 5) 265 34908A 40-Channel Multiplexer Component Locator 266 34908A 40-Channel Multiplexer Schematic (Sheet 1 of 3) 267 34908A 40-Channel Multiplexer Schematic (Sheet 2 of 3) 268 34908A 40-Channel Multiplexer Schematic (Sheet 3 of 3) 269
Contents
14
1
1
DC, Resistance, and Temperature Accuracy Specifications, on page 16 DC Measurement and Operating Characteristics, on page 17 AC Accuracy Specifications, on page 18 AC Measurement and Operating Characteristics, on page 19 Measurement Rates and System Characteristics, on page 20 Module Specifications: 34901A, 34902A, 34908A, 34903A, 34904A, on page 21 34905A, 34906A, on page 22 Typical AC Performance Graphs, on page 23 34907A, on page 24 BenchLink Data Logger Software Specifications, on page 24 Product and Module Dimensions, on page 25 To Calculate Total Measurement Error, on page 26 Interpreting Multimeter Specifications, on page 28 Configuring for Highest Accuracy Measurements, on page 31
Specifications
DC Voltage
100.0000 mV 1.000000 V 10.00000 V 100.0000 V 300.000 V 100.0000 1.000000 k 10.00000 k 100.0000 k 1.000000 M 10.00000 M 100.0000 M 10.00000 mA 100.0000 mA 1.000000 A Type
[6]
Resistance [4]
DC Current
34901A Only
Temperature Thermocouple
Best Range Accuracy [5] 1100C to 1820C -150C to 1000C -150C to 1200C -100C to 1200C -100C to 1300C 300C to 1760C 400C to 1760C -100C to 400C -200C to 600C -80C to 150C 1.2C 1.0C 1.0C 1.0C 1.0C 1.2C 1.2C 1.0C 0.06C 0.08C
Extended Range Accuracy [5] 400C to 1100C -200C to -150C -210C to -150C -200C to -100C -200C to -100C -50C to 300C -50C to 400C -200C to -100C 1.8C 1.5C 1.2C 1.5C 1.5C 1.8C 1.8C 1.5C 0.03C 0.03C 0.03C 0.03C 0.03C 0.03C 0.03C 0.03C 0.003C 0.002C
RTD Thermistor
[1] Specifications are for 1 hour warm up and 61 2 digits [2] Relative to calibration standards [3] 20% over range on all ranges except 300 Vdc and 1 Adc ranges [4] Specifications are for 4-wire ohms function or 2-wire ohms using Scaling to remove the offset. Without Scaling, add 1 additional error in 2-wire ohms function. [5] 1 year accuracy. For total measurement accuracy, add temperature probe error. [6] Thermocouple specifications not guaranteed when 34907A module is present
16
1
DC Measurement and Operating Characteristics
DC Measurement Characteristics [1]
DC Voltage Measurement Method: A/D Linearity: Input Resistance: 100 mV, 1 V, 10 V ranges 100 V, 300 V ranges Input Bias Current: Input Protection: Resistance Measurement Method: Offset Compensation: Max. Lead Resistance: Input Protection: DC Current Shunt Resistance: Input Protection: Thermocouple Conversion: Reference Junction Type: Open T/C Check: RTD Thermistor DC CMRR: Integration Time 200 PLC / 3.33s (4s) 100 PLC / 1.67s (2s) 20 PLC / 333 ms (400 ms) 10 PLC / 167 ms (200 ms) 2 PLC / 33.3 ms (40 ms) 1 PLC / 16.7 ms (20 ms) < 1 PLC Continuously Integrating, Multi-slope III A/D Converter 0.0002% of reading + 0.0001% of range Selectable 10 M or >10 G 10 M 1% < 30 pA at 25 C 300 V on all ranges Selectable 4-wire or 2-wire Ohms, Current source reference to LO input Selectable on 100, 1 k, 10 k ranges 10% of range per lead for 100 and 1 k ranges. 1 k on all other ranges 300 V on all ranges 5 for 10 mA, 100 mA; 0.1 for 1A. 1.5A 250 V fuse on 34901A module ITS-90 software compensation Internal, Fixed, or External Selectable per channel. Open > 5 k
= 0.00385 (DIN) and 0.00391
Single Channel Measurement Rates [8] Function DCV, 2-Wire Ohms: Resolution 61 2 (10 PLC) 51 2 (1 PLC) 41 2 (0.02 PLC) 0.1 C (1 PLC) 0.1 C (0.02 PLC) 0.01 C (10 PLC) 0.1 C (1 PLC) 1 C (0.02 PLC)
Autozero OFF Operation Following instrument warm-up at calibration temperature 1 C and < 10 minutes, add 0.0002% range additional error + 5 V. Settling Considerations Reading settling times are affected by source impedance, low dielectric absorption characteristics, and input signal changes.
44004, 44007, 44006 series 140 dB Normal Mode Rejection [3] 110 dB [4] 105 dB [4] 100 dB [4] 95 dB [4] 90 dB 60 dB 0 dB
Measurement Noise Rejection 60 Hz (50 Hz) [2] [1] 300 Vdc isolation voltage (ch-ch, ch-earth) [2] For 1 k unbalance in LO lead [3] For power line frequency 0.1% [4] For power line frequency 1%, use 80 dB. For power line frequency 3%, use 60 dB. [5] Reading speeds for 60 Hz and (50 Hz) operation; autozero OFF [6] 61 2 digits=22 bits, 51 2 digits=18 bits, 41 2 digits=15 bits [7] Add 20 V for DCV, 4 A for DCI, or 20 m for resistance [8] For fixed function and range, readings to memory, scaling and alarms off, autozero OFF
17
AC Accuracy Specifications
( % of reading + % of range ) [1]
Function Range [3] Includes measurement error, switching error, and transducer conversion error Frequency 24 Hour [2] 23 C 1 C 1.00 + 0.03 0.35 + 0.03 0.04 + 0.03 0.10 + 0.05 0.55 + 0.08 4.00 + 0.50 1.00 + 0.05 0.35 + 0.05 0.04 + 0.05 0.10 + 0.10 0.55 + 0.20 4.00 + 1.25 0.10 0.05 0.03 0.006 1.00 + 0.04 0.30 + 0.04 0.10 + 0.04 1.00 + 0.5 0.30 + 0.5 0.10 + 0.5 90 Day 23 C 5 C 1.00 + 0.04 0.35 + 0.04 0.05 + 0.04 0.11 + 0.05 0.60 + 0.08 4.00 + 0.50 1.00 + 0.08 0.35 + 0.08 0.05 + 0.08 0.11 + 0.12 0.60 + 0.20 4.00 + 1.25 0.10 0.05 0.03 0.01 1.00 + 0.04 0.30 + 0.04 0.10 + 0.04 1.00 + 0.5 0.30 + 0.5 0.10 + 0.5 1 Year 23 C 5 C 1.00 + 0.04 0.35 + 0.04 0.06 + 0.04 0.12 + 0.05 0.60 + 0.08 4.00 + 0.50 1.00 + 0.08 0.35 + 0.08 0.06 + 0.08 0.12 + 0.12 0.60 + 0.20 4.00 + 1.25 0.10 0.05 0.03 0.01 1.00 + 0.04 0.30 + 0.04 0.10 + 0.04 1.00 + 0.5 0.30 + 0.5 0.10 + 0.5 Temperature Coefficient /C 0 C 18 C 28 C 55 C 0.100 + 0.004 0.035 + 0.004 0.005 + 0.004 0.011 + 0.005 0.060 + 0.008 0.20 + 0.02 0.100 + 0.008 0.035 + 0.008 0.005 + 0.008 0.011 + 0.012 0.060 + 0.020 0.20 + 0.05 0.005 0.005 0.001 0.001 0.100 + 0.006 0.035 + 0.006 0.015 + 0.006 0.100 + 0.06 0.035 + 0.06 0.015 + 0.06
100.0000 mV to 100 V
3 Hz 5 Hz 5 Hz 10 Hz 10 Hz 20 kHz 20 kHz 50 kHz 50 kHz 100 kHz 100 kHz 300 kHz [5] 3 Hz 5 Hz 5 Hz 10 Hz 10 Hz 20 kHz 20 kHz 50 kHz 50 kHz 100 kHz 100 kHz 300 kHz [5] 3 Hz 5 Hz 5 Hz 10 Hz 10 Hz 40 Hz 40 Hz 300 kHz 3 Hz 5 Hz 5 Hz 10 Hz 10 Hz 5 kHz
300.0000 V
Additional Low Frequency Error for ACV, ACI (% of reading) Frequency 10 Hz - 20 Hz 20 Hz - 40 Hz 40 Hz - 100 Hz 100 Hz - 200 Hz 200 Hz - 1 kHz > 1 kHz AC Filter Slow 0 0 0 0 0 0 AC Filter Medium 0.74 0.22 0.06 0.01 0 0 AC Filter Fast 0.73 0.22 0.18 0
Additional Error for Frequency, Period (% of reading) Frequency 3 Hz - 5 Hz 5 Hz - 10 Hz 10 Hz - 40 Hz 40 Hz - 100 Hz 100 Hz - 300 Hz 300 Hz - 1 kHz > 1 kHz 612 Digits 0 0 0 0 0 0 0 512 Digits 0.12 0.17 0.2 0.06 0.03 0.01 0 412 Digits 0.12 0.17 0.2 0.21 0.21 0.07 0.02
[1] Specifications are for 1 hour warm up and 61 2 digits, Slow ac filter [2] Relative to calibration standards [3] 20% over range on all ranges except 300 Vac and 1 Aac ranges [4] For sinewave input > 5% of range. For inputs from 1% to 5% of range and < 50 kHz, add 0.1% of range additional error. [5] Typically 30% of reading error at 1 MHz, limited to 1x108 V Hz [6] Input > 100 mV. For 10 mV to 100 mV inputs, multiply % of reading error x 10. [7] Specified only for inputs > 10 mA
18
1
AC Measurement and Operating Characteristics
AC Measurement Characteristics [1]
True RMS AC Voltage Measurement Method: Crest Factor: Additional Crest Factor Errors (non-sinewave): [2] AC-coupled True RMS measures the ac component of input with up to 300 Vdc of bias on any range Maximum 5:1 at Full Scale Crest Factor 1-2: Crest Factor 2-3: Crest Factor 3-4: Crest Factor 4-5: 0.05% of reading 0.15% of reading 0.30% of reading 0.40% of reading
Single Channel Measurement Rates [8] Function ACV: Resolution 61 2 Slow (3 Hz) 61 2 Medium (20 Hz) 61 2 Fast (200 Hz) 6 1/2 [7] 61 2 Digits (1s gate) 51 2 Digits (100 ms) 41 2 Digits (10 ms)
AC Filter Bandwidth: Slow Medium Fast Input Impedance: Input Protection: Frequency and Period Measurement Method: Voltage Ranges: Gate Time: Measurement Timeout: True RMS AC Current Measurement Method:
3 Hz 300 kHz 20 Hz 300 kHz 200 Hz 300 kHz 1 M 2%, in parallel with 150 pF 300 Vrms on all ranges Reciprocal counting technique Same as AC Voltage function 1s, 100 ms, or 10 ms Selectable 3 Hz, 20 Hz, 200 Hz LF limit Direct coupled to the fuse and shunt. AC-coupled True RMS measurement (measures the ac component only) 5 for 10 mA; 0.1 for 100 mA, 1A 1.5A 250 V fuse on 34901A module 70 dB
Frequency, Period:
Measurement Noise Rejection [3] Measurement Considerations (Frequency and Period) All frequency counters are susceptible to error when measuring low-voltage, low-frequency signals. Shielding inputs from external noise pickup is critical for minimizing measurement errors.
[1] 300 Vrms isolation voltage (ch-ch, ch-earth) [2] For frequencies below 100 Hz, slow AC filter specified for sinewave input only [3] For 1 k unbalance in LO lead [4] Maximum reading rates for 0.01% of ac step additional error. Additional settling delay required when input dc level varies. [5] 61 2 digits=22 bits, 51 2 digits=18 bits, 41 2 digits=15 bits [6] For external trigger or remote operation using default settling delay (Delay Auto) [7] Maximum limit with default settling delays defeated [8] For fixed function and range, readings to memory, scaling and alarms turned off
19
System Characteristics
Readings/s 6 (5) 53 (47) 490 49 (47) 280 6 (5) 47 (47) 280 0.14 1 8 100 1 9 70 Scan Triggering Scan Count: Scan Interval: Channel Delay: External Trig Delay: External Trig Jitter: Alarms Alarm Outputs: Latency: Memory Readings: Time Stamp Resolution: Relative Absolute States: Alarm Queue: General Specifications Power Supply: Power Line Frequency: Power Consumption: Operating Environment: Storage Environment: Weight (Mainframe): Safety: RFI and ESD: Warranty: 1 to 50,000 or continuous 0 to 99 hours; 1 ms step size 0 to 60 seconds/channel; 1 ms step size < 300 s; With Monitor On, < 200 ms < 2 ms 4 TTL compatible. Selectable TTL logic HI or LO on Fail 5 ms (typical) Battery Backed, 4 year typical life [6] 50,000 readings 1 ms 1s 5 instrument states Up to 20 events 100 V / 120 V / 220 V / 240 V 10% 45 Hz to 66 Hz automatically sensed (12 W) 25 VA peak Full accuracy for 0 C to 55 C Full accuracy to 80% R.H. at 40 C -40 C to 70 C [6] Net: 3.6 kg (8.0 lbs) IEC 61010-1:2001 / EN 61010-1:2001 Canada: CSA C22.2 No. 61010-1:2004 UL 61010-1:2004 CISPR 11, IEC 801/2/3/4 1 year
ACV:
Frequency, Period:
INTO and OUT of Memory to GPIB or RS-232 (INIT, FETCh) 34902A Scanning DCV 180 34902A Scanning DCV with Time stamp 150 OUT of Memory to GPIB Readings Readings with Time stamp Readings with all Format Options ON OUT of Memory to RS-232 Readings Readings with Time stamp Readings with all Format Options ON DIRECT to GPIB or RS-232 Single Channel DCV 34902A Scanning DCV Single Channel MEAS DCV 10 or MEAS DCV 1 Single Channel MEAS DCV or MEAS OHMS
[5]
[1] Reading speeds for 60 Hz and (50 Hz) operation; autozero OFF [2] For fixed function and range, readings to memory, scaling and alarms off, autozero OFF [3] Maximum limit with default settling delays defeated [4] Speeds are for 41 2 digits, delay 0, display off, autozero off. Using 115 kbaud RS-232 setting. [5] Assumes relative time format (time since start of scan) [6] Storage at temperatures above 40 C will decrease battery life
This ISM device complies with Canadian ICES-001. Cet appareil ISM est conforme la norme NMB-001 du Canada.
N10149
20
1
Module Specifications
34901A, 34902A, 34908A, 34903A, 34904A
Multiplexer General Number of Channels Connects to Internal DMM Scanning Speed [1] Open/Close Speed Maximum Input Voltage (dc, ac rms) Current (dc, ac rms) Power (W, VA) Isolation (ch-ch, ch-earth) DC Characteristics Offset Voltage [2] Initial Closed Channel R [2] Isolation (ch-ch, ch-earth) AC Characteristics Bandwidth Ch-Ch Cross Talk (dB) [3] Capacitance Capacitance Volt-Hertz Limit Other T/C Cold Junction Accuracy [2] [5] Switch Life
[6]
Actuator 34908A 40 1 wire Yes 60 ch/s 70/s 300 V 1A 50 W 300 V < 3 V < 1 > 10 G 10 MHz -18 [4] < 50 pF < 80 pF 108 0.8 C [7] 100M 100k 100M 100k 120/s 300 V 1A 50 W 300 V < 3 V < 0.2 > 10 G 10 MHz -45 < 10 pF < 80 pF 108 34903A 20 SPDT No
Matrix 34904A 4x8 2 wire No 120/s 300 V 1A 50 W 300 V < 3 V < 1 > 10 G 10 MHz -33 < 50 pF < 80 pF 108
34901A 20+2 2/4 wire Yes 60 ch/s 120/s 300 V 1A 50 W dc, ac rms 300 V < 3 V < 1 > 10 G 10 MHz 10 MHz HI to LO LO to Earth -45 < 50 pF < 80 pF 108 (typical) 0.8 C 100M 100k
34902A 16 2/4 wire Yes 250 ch/s 120/s 300 V 50 mA 2W 300 V < 6 V < 1 > 10 G 10 MHz -45 < 50 pF < 80 pF 108 0.8 C 100M 100k
100M 100k
Humidity (non-condensing)
[1] Speeds are for 41 2 digits, delay 0, display off, autozero off. Using 115 kbaud RS-232 setting. [2] Errors included in the DMM measurement accuracy specifications [3] 50 source, 50 load [4] Isolation within channel 1 to 20 or 21 to 40 banks is -40 dB [5] Thermocouple specifications not guaranteed when 34907A module is present [6] Applies to resistive loads only [7] Thermocouple measurements not recommended with 34908A module due to common LO configuration.
21
Module Specifications
34905A, 34906A
RF Multiplexer General Number of Channels Open/Close Speed Maximum Input Voltage (dc, ac rms) Current (dc, ac rms) Power (W, VA) DC Characteristics Offset Voltage
[1]
Initial Closed Channel R [1] Isolation (ch-ch, ch-earth) Other Switch Life Switch Life Temperature Temperature Humidity (non-condensing) No Load (typical) Rated Load (typical) [2] Operating Storage
[1] Errors included in DMM measurement accuracy specifications [2] Applies to resistive loads only [3] Bandwidth direct to module SMB connectors [4] 50 ohm source, 50 ohm load
22
1
Typical AC Performance Graphs
34905A, 34906A
Insertion Loss (50) Insertion Loss (75)
VSWR (50)
VSWR (75)
Crosstalk (50)
Crosstalk (75)
23
Module Specifications
34907A
Digital Input / Output
Port 1, 2: Vin(L): Vin(H): Vout(L): Vout(H): Vin(H) Max: Alarming: Speed Latency Read/Write Speed: 8 Bit, input or output, non-isolated < 0.8V (TTL) > 2.0V (TTL) < 0.8V @ Iout = - 400 mA > 2.4V @ Iout = 1 mA < 42V with external open drain pull-up Maskable pattern match or state change 4 ms (max) alarm sampling 5 ms (typical) to 34970A alarm output 95/s
Software Specifications
BenchLink Data Logger (not included with Option 001)
System Requirements [1] 486, 66 MHz, 16 MB RAM, PC Hardware: 12 MB disk space Windows 3.1, Windows 95, Operating System: Windows NT 4.0 Computer Interfaces [2] Agilent 82335B, 82340A/B/C, GPIB: 82341A/B/C/D National Instruments AT-GPIB/TNT, PCI-GPIB LAN-to-GPIB: Agilent E5810A (Windows 98/Me/ NT/2000/XP Professional) RS-232 (Serial Port): PC COM 1 to 4 [3] Performance Scan and Save to Disk: 100 ch/s, 2 strip charts displayed [1] Software provided on CD-ROM; includes utility to create floppy disks for installation [2] Interface and drivers must be purchased and installed separately [3] 90 MHz Pentium, 20 MB RAM
Totalize Input
Maximum Count: Totalize Input: Signal Level: Threshold: Gate Input: Count Reset: Read Speed: 226 - 1 (67,108,863) 100 kHz (max), rising or falling edge, programmable 1 Vp-p (min) 42 Vpk (max) 0V or TTL, jumper selectable TTL-Hi, TTL-Lo, or none Manual or Read+Reset 85/s
[1] Limited to 40 mA total for all three slots (six DAC channels)
24
1
Product and Module Dimensions
103.6 mm
254.4 mm
374.0 mm
88.5 mm
212.6 mm
348.3 mm
Module
TOP
315.6
91.9
25
26
1
Understanding the % of range Error The range error compensates for inaccuracies that result from the function and range you select. The range error contributes a constant error, expressed as a percent of range, independent of the input signal level. The following table shows the range error applied to the DMMs 24-hour dc voltage specification.
Range Error (% of range) 0.0004 0.0004 0.0004 Range Error Voltage
40 V 40 V 40 V
Total Measurement Error To compute the total measurement error, add the reading error and range error. You can then convert the total measurement error to a percent of input error or a ppm (part-permillion) of input error as shown below. % of input error =
Total Measurement Error 100 Input Signal Level Total Measurement Error 1,000,000 Input Signal Level
Assume that a 5 Vdc signal is input to the DMM on the 10 Vdc range. Compute the total measurement error using the 90-day accuracy specification of (0.0020% of reading + 0.0005% of range). Reading Error Range Error Total Error = 0.0020% x 5 Vdc = 0.0005% x 10 Vdc = 100 V + 50 V = 100 V = 50 V = 150 V = 0.0030% of 5 Vdc = 30 ppm of 5 Vdc
27
Sensitivity
Sensitivity is the minimum level that the multimeter can detect for a given measurement. Sensitivity defines the ability of the multimeter to respond to small changes in the input level. For example, suppose you are monitoring a 1 mVdc signal and you want to adjust the level to within 1 V. To be able to respond to an adjustment this small, this measurement would require a multimeter with a sensitivity of at least 1 V. You could use a 612-digit multimeter if it has a 1 Vdc or smaller range. You could also use a 412-digit multimeter with a 10 mVdc range. For ac voltage and ac current measurements, note that the smallest value that can be measured is different from the sensitivity. For the internal DMM, these functions are specified to measure down to 1% of the selected range. For example, the internal DMM can measure down to 1 mV on the 100 mV range.
28
1
Resolution
Resolution is the numeric ratio of the maximum displayed value divided by the minimum displayed value on a selected range. Resolution is often expressed in percent, parts-per-million (ppm), counts, or bits. For example, a 612-digit multimeter with 20% overrange capability can display a measurement with up to 1,200,000 counts of resolution. This corresponds to about 0.0001% (1 ppm) of full scale, or 21 bits including the sign bit. All four specifications are equivalent.
Accuracy
Accuracy is a measure of the exactness to which the internal DMMs measurement uncertainty can be determined relative to the calibration reference used. Absolute accuracy includes the Internal DMMs relative accuracy specification plus the known error of the calibration reference relative to national standards (such as the U.S. National Institute of Standards and Technology). To be meaningful, the accuracy specifications must be accompanied with the conditions under which they are valid. These conditions should include temperature, humidity, and time. There is no standard convention among multimeter manufacturers for the confidence limits at which specifications are set. The table below shows the probability of non-conformance for each specification with the given assumptions.
Specification Criteria Mean 2 sigma Mean 3 sigma Probability of Failure 4.5% 0.3%
Variations in performance from reading to reading, and instrument to instrument, decrease for increasing number of sigma for a given specification. This means that you can achieve greater actual measurement precision for a specific accuracy specification number. The 34970A is designed and tested to meet performance better than mean 3 sigma of the published accuracy specifications.
29
24-Hour Accuracy
The 24-hour accuracy specification indicates the internal DMMs relative accuracy over its full measurement range for short time intervals and within a stable environment. Short-term accuracy is usually specified for a 24-hour period and for a 1 C temperature range.
Temperature Coefficients
Accuracy is usually specified for a 23 C 5 C temperature range. This is a common temperature range for many operating environments. You must add additional temperature coefficient errors to the accuracy specification if you are operating the multimeter outside a 23 C 5 C temperature range (the specification is per C).
30
1
Configuring for Highest Accuracy Measurements
The measurement configurations shown below assume that the internal DMM is in its Factory Reset state. It is also assumed that manual ranging is enabled to ensure proper full scale range selection. DC Voltage, DC Current, and Resistance Measurements: Set the resolution to 6 digits (you can use the 6 digits slow mode for further noise reduction). Set the input resistance to greater than 10 G (for the 100 mV, 1 V, and 10 V ranges) for the best dc voltage accuracy. Use 4-wire ohms and enable offset compensation for the best resistance accuracy. AC Voltage and AC Current Measurements: Set the resolution to 6 digits. Select the slow ac filter (3 Hz to 300 kHz). Frequency and Period Measurements: Set the resolution to 6 digits.
31
32
Quick Start
Quick Start
One of the first things you will want to do with your instrument is to become acquainted with the front panel. We have written the exercises in this chapter to prepare the instrument for use and help you get familiar with some of its front-panel operations. The front panel has several groups of keys to select various functions and operations. A few keys have a shifted function printed in blue below the key. To perform a shifted function, press (the SHIFT annunciator will turn on). Then, press the key that has the desired label below it. For example, to select the Utility Menu, press . If you accidentally press annunciator. , just press it again to turn off the SHIFT
This chapter is divided into the following sections: To Prepare the Instrument for Use, on page 35 To Connect Wiring to a Module, on page 36 To Set the Time and Date, on page 38 To Configure a Measurement Channel, on page 39 To Monitor a Single Channel, on page 40 To Close a Channel, on page 41 If the Instrument Does Not Turn On, on page 42 To Adjust the Carrying Handle, on page 44 To Rack Mount the Instrument, on page 45
34
2
1 Check the list of supplied items. Verify that you have received the following items with your instrument. If anything is missing, contact your nearest Agilent Technologies Sales Office. One power cord. One Users Guide. This Service Guide. One Quick Reference Guide. Certificate of Calibration (if you ordered the internal DMM). Quick Start Kit (if you ordered the internal DMM): One RS-232 cable. BenchLink Data Logger Software CD-ROM. One J-type thermocouple and a flatblade screwdriver. Any plug-in modules that you ordered are delivered in a separate shipping container.
On/Standby Switch WARNING Note that this switch is Standby only. To disconnect the mains from the instrument, remove the power cord.
2 Connect the power cord and turn on the instrument. The front-panel display will light up briefly while the instrument performs its power-on self-test. The GPIB address is displayed. The instrument initially powers up with all measurement channels turned off. To review the power-on display with all annunciators turned on, hold down as you turn on the instrument. If the instrument does not turn on properly, see page 42. 3 Perform a complete self-test. The complete self-test performs a more extensive set of tests than those performed at power-on. Hold down as you turn on the instrument and hold down the key until you hear a long beep. The self-test will begin when you release the key following the beep.
35
20 AWG Typical 6 mm
Wiring Hints...
For detailed information on each module, refer to the 34970A Users Guide. To reduce wear on the internal DMM relays, wire like functions on adjacent channels. Use shielded twisted pair PTFE insulated cables to reduce settling and noise errors. The diagrams on the next page show how to connect wiring to a multiplexer module for each measurement function.
36
Thermocouple
Thermocouple Types: B, E, J, K, N, R, S, T
Ranges: 100, 1 k, 10 k, 100 k, 1 M, 10 M, 100 M RTD Types: 0.00385, 0.00391 Thermistor Types: 2.2 k, 5 k, 10 k
DC Current / AC Current
Channel n (source) is automatically paired with Channel n+10 (sense) on the 34901A or Channel n+8 (sense) on the 34902A. Valid only on channels 21 and 22 on the 34901A. Ranges: 10 mA, 100 mA, 1A Ranges: 100, 1 k, 10 k, 100 k, 1 M, 10 M, 100 M RTD Types: 0.00385, 0.00391
37
Use and to select the field to modify and turn the knob to change the value. You can also edit the AM/PM field. TIME 03:45 PM
JUNE 01 2002
38
2 Select the measurement parameters for the selected channel. Use the knob to scroll through the measurement choices on each level of the menu. When you press to make your selection, the menu automatically guides you through all relevant choices to configure a measurement on the selected function. When you have finished configuring the parameters, you are automatically exited from the menu. The present selection (or default) is displayed in full bright for easy identification. When you make a different selection, the new choice is shown in full bright and it becomes the default selection. The order of the choices always remains the same; however, you always enter the menu at the present (full-bright) setting for each parameter. Note: The menu will timeout after about 20 seconds of inactivity and any changes made previously will take effect.
39
40
To Close a Channel
On the multiplexer and switch modules, you can close and open individual relays on the module. However, note that if you have already configured any multiplexer channels for scanning, you cannot independently close and open individual relays on that module. 1 Select the channel. Turn the knob until the desired channel is shown on the right side of front-panel display. For this example, select channel 213. 2 Close the selected channel. 3 Open the selected channel. Note: will sequentially open all channels on the module in the selected slot. The table below shows the low-level control operations available for each of the plug-in modules.
Plug-In Module 34901A 34902A 34908A 34903A 34904A 34905A 34906A 34907A 34907A 34907A 20-Channel Mux 16-Channel Mux 40-Channel Single-Ended Mux [1] 20-Channel Actuator 4x8 Matrix Dual 4-Channel RF Mux (50 ) [2] Dual 4-Channel RF Mux (75) [2] Multifunction Module (DIO) Multifunction Module (Totalizer) Multifunction Module (DAC) ,
[1] Only one channel can be closed at a time on this module. [2] Only one channel in each bank can be closed at a time on this module.
41
2 Verify the power-line voltage setting. The line voltage is set to the proper value for your country when the instrument is shipped from the factory. Change the voltage setting if it is not correct. The settings are: 100, 120, 220, or 240 Vac. Note: For 127 Vac operation, use the 120 Vac setting. For 230 Vac operation, use the 220 Vac setting. See the next page if you need to change the line-voltage setting. 3 Verify that the power-line fuse is good. The instrument is shipped from the factory with a 500 mA fuse installed. This is the correct fuse for all line voltages. See the next page if you need to replace the power-line fuse.
To replace the 500 mAT, 250 V fuse, order Agilent part number 2110-0458.
42
1 Remove the power cord. Remove the fuse-holder assembly from the rear panel.
Fuse: 500 mAT (for all line voltages) Agilent Part Number: 2110-0458
Verify that the correct line voltage is selected and the power-line fuse is good.
43
Carrying Position
44
To remove the handle, rotate it to the vertical position and pull the ends outward.
Front
To remove the rubber bumper, stretch a corner and then slide it off.
45
To rack mount two instruments side-by-side, order lock-link kit 5061-9694 and flange kit 5063-9212. Be sure to use the support rails inside the rack cabinet.
To install one or two instruments in a sliding support shelf, order shelf 5063-9255, and slide kit 1494-0015 (for a single instrument, also order filler panel 5002-3999).
46
Front-Panel Overview
Front-Panel Overview
This chapter introduces you to the front-panel keys and menu operation. This chapter does not give a detailed description of every front-panel key or menu operation. It does, however, give you a good overview of the front-panel menu and many front-panel operations. See the Agilent 34970A Users Guide for a complete discussion of the instruments capabilities and operation. This chapter is divided into the following sections: Front-Panel Menu Reference, on page 49 To Unsecure for Calibration, on page 51 To Secure Against Calibration, on page 51 To Change the Security Code, on page 52 Error Messages, on page 52 To Perform a Zero Adjustment, on page 53 To Apply Mx+B Scaling to Measurements, on page 54 To Read the Relay Cycle Count, on page 55 To Read a Digital Input Port, on page 56 To Write to a Digital Output Port, on page 57 To Read the Totalizer Count, on page 58 To Output a DC Voltage, on page 59
48
49
50
When you first enter the Utility menu, the calibration entries toggle between CAL SECURED and UNSECURE CAL. To unsecure the instrument, select UNSECURE CAL and press . After entering the correct security code, press again. When you return to the menu, you will see new choices CAL UNSECURED and SECURE CAL. Note: If you enter the wrong secure code, NO MATCH is displayed and a new choice, EXIT, is shown.
When you enter the Utility menu, the calibration entries toggle between CAL UNSECURED and SECURE CAL. To secure the instrument, select SECURE CAL and press . After entering the desired security code, press again. When you return to the menu, you will see new choices CAL SECURED and UNSECURE CAL.
51
Error Messages
Error messages are retrieved in a first-in first-out (FIFO) order. When the ERROR annunciator is on, press to view error messages. Use the arrow keys to scroll the message in the display. A list of the self-test errors messages and their meanings begin on page 168. For a complete list of error messages and descriptions, see chapter 6 in the 34970A Users Guide.
52
5 Begin the adjustment. The display will show the progress of the adjustment. When all the adjustments are completed, the display will show done. DONE
53
+1.000,000
-0.700,000
OHM
3 Select the custom label. You can specify an optional three-character label for your scaled readings (RPM, PSI, etc.). The default label is the standard engineering unit for the selected function (VDC, OHM, etc.). LABEL AS OHM
54
RELAY CYCLES
55
The bit pattern read from the port will be displayed until you press another key, turn the knob, or until the display times out. Note: To add a digital input channel to a scan list, press the DIO READ choice. and select
56
3 Edit the bit pattern. Use the knob and or keys to edit the individual bit values. You can specify whether you want to use binary or decimal format. Once you have selected the number base, it is used for all input or output operations on the same port. To change the number base, press the key and select USE BINARY or USE DECIMAL. 240 DOUT Decimal Display Shown
4 Output the bit pattern to the specified port. The specified bit pattern is latched on the specified port. To cancel an output operation in progress, wait for the display to time out.
57
3 Read the count. The count is read once each time you press ; the count does not update automatically on the display. As configured in this example, the count is automatically reset to 0 each time you read it. 12345 TOT
The count will be displayed until you press another key, turn the knob, or until the display times out. To manually reset the totalizer count, press . Note: To add a totalizer channel to a scan list, press and select the
58
To Output a DC Voltage
The multifunction module (34907A) has two analog outputs capable of outputting calibrated voltages between 12 volts. 1 Select a DAC Output channel. Select the slot containing the multifunction module and continue turning the knob until DAC is displayed (channel 04 or 05). 2 Enter the output voltage editor. +00.000 V DAC
3 Set the desired output voltage. Use the knob and or keys to edit the individual digits.
+05.250 V DAC
4 Output the voltage from the selected DAC . The output voltage will be displayed until you press another key or turn the knob. To manually reset the output voltage to 0 volts, press .
59
60
Calibration Procedures
Calibration Procedures
This chapter contains procedures for verification of the instruments performance and adjustment (calibration). These procedures are required only if the internal DMM is installed. The chapter is divided into the following sections: Agilent Technologies Calibration Services, on page 63 Calibration Interval, on page 63 Time Required for Calibration, on page 64 Automating Calibration Procedures, on page 64 Recommended Test Equipment, on page 65 Input connections, on page 66 Calibration Security, on page 67 Calibration Message, on page 69 Calibration Count, on page 69 Calibration Procedures, on page 70 Aborting a Calibration in Progress, on page 70 Test Considerations, on page 71 Performance Verification Tests, on page 72 Internal DMM Verification Tests, on page 75 Optional AC Performance Verification Tests, on page 80 Internal DMM Adjustments, on page 81 10 Vdc Adjustment Procedure (Optional), on page 85 Plug-in Module Test Considerations, on page 87 Relay Verification, on page 88 Thermocouple Reference Junction (Optional), on page 112 34907A Analog Output, on page 114
Closed-Case Electronic Calibration The instrument features closed-case electronic calibration. No internal mechanical adjustments are required. The instrument calculates correction factors based upon the input reference value you set. The new correction factors are stored in non-volatile memory until the next calibration adjustment is performed. Non-volatile EEPROM calibration memory does not change when power has been off or after a remote interface reset.
62
Calibration Interval
The instrument should be calibrated on a regular interval determined by the measurement accuracy requirements of your application. A 1-year interval is adequate for most applications. Accuracy specifications are warranted only if adjustment is made at regular calibration intervals. Accuracy specifications are not warranted beyond the 1-year calibration interval. Agilent does not recommend extending calibration intervals beyond 2 years for any application.
Adjustment is Recommended
Whatever calibration interval you select, Agilent recommends that complete re-adjustment should always be performed at the calibration interval. This will assure that the 34970A will remain within specification for the next calibration interval. This criteria for re-adjustment provides the best long-term stability. Performance data measured using this method can be used to extend future calibration intervals. Use the Calibration Count feature (see page 69) to verify that all adjustments have been performed.
63
64
Recommended Equipment None Fluke 5700A Fluke 5700A/ 5725A Fluke 5700A Fluke 5700A/ 5725A Fluke 5700A/ 5725A Agilent 33220A Agilent 34401A Thermistor YSI 44031 (two) [2] J Type Calibrated Thermocouple Triple Point Cell Agilent 34401A
Accuracy Requirements 4 -terminal all copper short <1/5 instrument 24 hour spec <1/5 instrument 24 hour spec <1/5 instrument 24 hour spec <1/5 instrument 24 hour spec <1/5 instrument 24 hour spec <1/5 instrument 24 hour spec <1/5 instrument 24 hour spec 0.1 C
0.001 resolution
[1] In addition to the internal DMM, these applications require an input multiplexer module. The 34901A is recommended. [2] Thermistor YSI 44031 is available as Agilent part number 34308A (package of five).
65
Input Connections
You will need an input multiplexer module to verify or adjust the internal DMM. Input connections can be made using a 34901A 20-Channel Multiplexer. To use a 34901A to completely verify and adjust the internal DMM, make the following connections:
Connections for DC/AC Current Connections for 4-wire Ohms Connections for DC V, AC V, 2- and 4-wire Ohms
Copper Short
34901A
To Calibrator
Note: Use shielded twisted pair PTFE insulated cables to reduce settling and noise errors. Connect the shield to the source LO output. You can also use a 34902A for test and adjustment of voltage, frequency, and resistance functions. You cannot test or adjust current inputs with a 34902A. If you use a 34902A; connect the copper shorts to Channels 7 and 15 and make the input connections to Channels 8 and 16.
66
Calibration Security
This feature allows you to enter a security code to prevent accidental or unauthorized adjustments of the instrument. When you first receive your instrument, it is secured. Before you can adjust the instrument, you must unsecure it by entering the correct security code. See page 51 in Chapter 3 for a procedure to enter the security code. The security code is set to HP034970 when the instrument is shipped from the factory. The security code is stored in non-volatile memory, and does not change when power has been off, after a Factory Reset (*RST command), or after an Instrument Preset (SYSTem:PRESet command). The security code may contain up to 12 alphanumeric characters. The first character must be a letter, but the remaining characters can be letters, numbers, or an underscore ( _ ). You do not have to use all 12 characters but the first character must always be a letter. Note: If you forget your security code, you can disable the security feature by adding a jumper inside the instrument as described on the following page.
67
Disconnect the power cord and all input connections. Remove the instrument cover (see page 174). Turn the instrument over. Apply power and turn on the instrument. Be careful not to touch the power line connections. Apply a short between the two exposed metal pads marked CAL UNLOCK as shown in the figure below.
Apply Short
5 6 7
While maintaining the short, enter any unsecure code. The instrument is now unsecured. Remove the short. Turn off the instrument and remove the power cord. Reassemble the instrument. Now you can enter a new security code. Be sure to remember the new security code.
68
Calibration Message
The instrument allows you to store one message in calibration memory. For example, you can store such information as the date when the last calibration was performed, the date when the next calibration is due, the instruments serial number, or even the name and phone number of the person to contact for a new calibration. You can record a calibration message only from the remote interface and only when the instrument is unsecured. You can read the message from either the front-panel or over the remote interface. You can read the calibration message whether the instrument is secured or unsecured. The calibration message may contain up to 40 characters. From the front panel, you can view 13 characters of the message at a time. Press to scroll through the text of the message. Press again to increase the scrolling speed.
Calibration Count
You can query the instrument to determine how many calibrations have been performed. Note that your instrument was calibrated before it left the factory. When you receive your instrument, be sure to read the count to determine its initial value. The calibration count increments up to a maximum of 65,535 after which it rolls over to 0. Since the value increments by one for each calibration point, a complete calibration may increase the value by many counts. The calibration count is also incremented with calibrations of the DAC channels on the multifunction module. Front-Panel Operation: CAL COUNT Remote Interface Operation: CALibration:COUNt?
69
Calibration Procedure
The following procedure is the recommended method to complete an instrument calibration. 1 2 3 4 5 6 7 Read Test Considerations (page 71). Unsecure the instrument for calibration (page 51). Perform the verification tests to characterize the instrument (incoming data). Perform the zero adjustment procedures. Perform the gain adjustment procedures. Perform the verification tests to verify the adjustments (outgoing data). Secure the instrument against calibration. Note the new security code and calibration count in the instruments maintenance records.
70
Test Considerations
To ensure proper instrument operation, verify that you have selected the correct power line voltage prior to attempting any procedure in this chapter. See If the Instrument Does Not Turn On, on page 42. Errors may be induced by ac signals present on the input leads during a self-test. Long test leads can also act as an antenna causing pick-up of ac signals. For optimum performance, all procedures should comply with the following recommendations: Assure that the calibration ambient temperature is stable and between 18 C and 28 C. Ideally the calibration should be performed at 23 C 1 C. Assure ambient relative humidity is less than 80%. Allow a 2-hour warm-up period with a copper short connected and the multiplexer module installed before verification or adjustment. The connections are shown in the figure on page 66. Use shielded twisted pair PTFE insulated cables to reduce settling and noise errors. Keep the input cables as short as possible. Connect the input cable shield to the source LO output. Except where noted in the procedures, connect the calibrator LO source to earth ground. Two-wire Ohms measurements are affected by the entire path length, including the plug-in card trace length and slot trace lengths. On the 34901A, Channel 10 is recommended as the median path length (on the 34902A, use Channel 8) for 2-wire Ohms verification and adjustments. Install the input multiplexer in slot 200. Because the instrument is capable of making highly accurate measurements, you must take special care to ensure that the calibration standards and test procedures used do not introduce additional errors. Ideally, the standards used to verify and adjust the instrument should be an order of magnitude more accurate than each instrument range full scale error specification. For the dc voltage, dc current, and resistance gain verification measurements, you should take care to ensure the calibrators 0" output is correct. If necessary, the measurements can be referenced to the calibrators 0" output using Mx + B scaling (see page 54). You will need to set the offset for each range of the measuring function being verified.
71
72
Self-Test
A brief power-on self-test occurs automatically whenever you turn on the instrument. This limited test assures that the instrument is capable of operation and also checks the plug-in cards for basic operation. To perform a complete self-test hold down the key as you press the power switch to turn on the instrument; hold down the key for more than 5 seconds until the instrument beeps (a complete description of these tests can be found in chapter 6). The instrument will automatically perform the complete self-test procedure when you release the key. The self-test will complete in approximately 20 seconds. If the self-test is successful, PASS is displayed on the front panel. If the self-test fails, FAIL is displayed and the ERROR annunciator turns on. If repair is required, see chapter 6, Service, for further details. If all tests pass, you have a high confidence (~90%) that the instrument is operational.
73
74
75
Continued from previous page... 1 Select each function and range in the order shown in the table below. Before executing each test, you must press to enable reading monitoring on the selected channel (or use the ROUTe:MON command from the remote interface). Compare measurement results to the appropriate test limits shown in the table (see page 74).
Channel 221 Function[1] Range DC Current 10 mA 100 mA 1A Quick Check
Input Open Open Open Input Short Short Short Short Short Short Short Short Short Short Short Short
Q Quick Check
2 A 5 A 100 A
2 A 5 A 100 A
Channel 209 Range Function[1] DC Volts 100 mV 1V 10 V 100 V 300 V 100 1 k 10 k 100 k 1 M 10 M 100 M
Error from Nominal 24 hour 90 day 1 year 3.5 V 6 V 40 V 600 V 6 mV 3.5 m 6 m 50 m 500 m 10 100 10 k
4 V 7 V 50 V 600 V 9 mV
4 V 7 V 50 V 600 V 9 mV
4 m 10 m 100 m 1 10 100 10 k
4 m 10 m 100 m 1 10 100 10 k
[1] Select 612 digit resolution. [2] For 2-wire ohms, an additional 4 of error must be added. Q: Quick performance verification test points.
Note: Zero offset calibration using a multifunction calibrator is NOT recommended. The calibrator and cabling offset can be large and unstable causing poor offset calibration of the internal DMM.
76
Gain Verification
This procedure is used to check the full scale reading accuracy of the internal DMM. Verification checks are performed only for those functions and ranges with unique gain calibration constants. Begin verification by selecting a measuring function and range. Make sure you have read Test Considerations on page 71. DC VOLTS, Resistance, and DC CURRENT Gain Verification Test 1 2 Make sure you have read Test Considerations on page 71. Select each function and range in the order shown below. Before executing each test, you must press to enable reading monitoring on the selected channel (or use the ROUTe:MON command from the remote interface). Compare measurement results to the appropriate test limits shown in the table. (Be certain to allow for appropriate source settling.)
Channel 210 Quick Function[1] Range Check DC Volts 100 mV 1V Q 10 V 100 V Q 300 V 2-Wire Ohms[2] and 4-Wire Ohms 100 1 k 10 k 100 k 1 M 10 M 100 M Error from Nominal 24 hour 90 day 1 year 6.5 V 8 V 9 V 26 V 37 V 47 V 190 V 250 V 400 V 2.6 mV 4.1 mV 5.1 mV 12 mV 22.5 mV 19.5 mV 6.5 m 12 m 14 m 26 m 90 m 110 m 250 m 900 m 1.1 2.5 9 11 30 90 110 1.6 k 2.1 k 4.1 k 310 k 801 k 810 k Error from Nominal 24 hour 90 day 1 year 1.5 A 5 A 7 A 14 A 35 A 55 A 560 A 900 A 1.1 mA
Input 100 mV 1V 10 V 100 V 300 V 100 1 k 10 k 100 k 1 M 10 M 100 M [3] Input 10 mA 100 mA 1A
[1] Select 612 digit resolution. [2] The 2-wire ohms resistance verification test is optional (see note on Page 82). For 2-wire ohms, an additional 1 of error must be added (see Page 71). Add a 1-second channel delay when using Fluke 5700 in 2-wire compensated mode. This avoids response time issues with 2-wire compensation when 34970As current source contains a pulse. [3] Verify only, no adjustment required. Q: Quick performance verification test points.
77
AC VOLTS Gain Verification Test Configuration: 1 2 AC Volts LF 3 HZ:SLOW (in the Advanced menu)
Make sure you have read Test Considerations on page 71. Select Channel 210, set the AC VOLTS function and the 3 Hz input filter. With the slow filter selected, each measurement takes 7 seconds to complete. Before executing each test, you must press to enable reading monitoring on the selected channel (or use the ROUTe:MON command from the remote interface). Select each range in the order shown below. Compare measurement results to the appropriate test limits shown in the table. (Be certain to allow for appropriate source settling.)
Input Frequency 1 kHz 50 kHz 1 kHz 50 kHz 1 kHz 50 kHz 10 Hz 1 kHz 1 kHz 50 kHz 1 kHz 50 kHz Quick Check
Range 100 mV
Error from Nominal 24 hour 90 day 1 year 70 V 150 V 700 V 1.5 mV 7 mV 15 mV 7 mV 34 V 70 mV 150 mV 270 mV 600 mV
[1] For this test, isolate the calibrators output from earth ground. [2] Some calibrators may have difficulty driving the internal DMM and cable load at this V-Hz output. Use short, low capacitance cable to reduce calibration loading. Verification can be performed at >195 Vrms. New test limits can be computed from the accuracy specification shown in Chapter 1 for the actual test conditions used. Q: Quick performance verification test points.
Note: The 50 kHz ac voltage test points may fail performance verification if the internal shields have been removed and reinstalled. See Gain Adjustment, on page 82, for further information on how to recalibrate the ac voltage function.
78
AC CURRENT Gain Verification Test Configuration: 1 2 AC Current LF 3 HZ:SLOW (in the Advanced menu)
Make sure you have read Test Considerations on page 71. Select Channel 221, set the AC CURRENT function and the 3 Hz input filter. With the slow filter selected, each measurement takes 7 seconds to complete. Before executing each test, you must press to enable reading monitoring on the selected channel (or use the ROUTe:MON command from the remote interface). Select each range in the order shown below. Compare measurement results to the appropriate test limits shown in the table. (Be certain to allow for appropriate source settling.)
Input Current Frequency 10 mA [1] 1 kHz 1 kHz 100 mA [1] 1 kHz 10 mA 1 kHz 1A [1]
[1] Verify only, no adjustment.
Range 10 mA 100 mA 1A 1A
Quick check Q
Error from Nominal 24 hour 90 day 1 year 14 A 600 A 1.41 mA 1.4 mA 14 A 600 A 1.41 mA 1.4 mA 14 A 600 A 1.41 mA 1.4 mA
Make sure you have read Test Considerations on page 71. Select Channel 210, select the FREQUENCY function and set 612 digits. Select each range in the order shown below. Compare measurement results to the appropriate test limits shown in the table. (Be certain to allow for appropriate source settling.)
Voltage 10 mV [1] 1V Input Frequency 100 Hz 100 kHz Range 100 mV 1V Quick Check
0.06 Hz 6 Hz
0 .1 Hz 10 Hz
0 .1 Hz 10 Hz
[1] Verify only, No adjustment. For this test, isolate the calibrators output from earth ground. Q: Quick performance verification test points.
79
Make sure you have read Test Considerations on page 71. Select Channel 210, select the AC Volts function and the 3 HZ filter. Before executing each test, you must press to enable reading monitoring on the selected channel (or use the ROUTe:MON command from the remote interface). Select each range in the order shown below. Compare measurement results to the appropriate test limits shown in the table. (Be certain to allow for appropriate source settling.)
Voltage 1V 1V 1V 1V 10 V 1V 100 mV Input Frequency 20 Hz 20 kHz 100 kHz 300 kHz 1 kHz 1 kHz 1 kHz Range 1V 1V 1V 1V 10 V 10 V 10 V
Error from Nominal 24 hour 90 day 1 year 700 V 700 V 6.3 mV 45 mV 7 mV 3.4 mV 13 mV
1 mV 1 mV 6.8 mV 45 mV
9 mV 4.5 mV 14 mV
10 mV 4.6 mV 14 mV
80
Zero Adjustment
Each time you perform a zero adjustment, the Internal DMM stores a new set of offset correction constants for every measurement function and range. The Internal DMM will sequence through all required functions and ranges automatically and store new zero offset calibration constants. All offset corrections are determined automatically. You may not correct a single range or function without re-entering ALL zero offset correction constants automatically. This feature is intended to save calibration time and improve zero calibration consistency. Note: Never turn off the Internal DMM during Zero Adjustment. This may cause ALL calibration memory to be lost. Zero Adjustment Procedure The zero adjustment procedure takes about 5 minutes to complete. Be sure to allow the instrument to warm up for 2 hours before performing the adjustments. Follow the steps outlined below. Review Test Considerations on page 71 before beginning this test. Also see page 53, for an example of how to initiate a zero calibration. 1 2 3 4 5 This procedure will use the copper shorts installed on Channels 209 and 219. Leave the Amps input connections (Channel 221) open. Select Channel 209. Select the DC VOLTS function. Press to enter the calibration menu. Press begin the adjustment procedure. again to
Use the knob and arrow keys to set the number in the display to 0.000000 and press . Perform the Zero Offset Verification tests (see page 75) to check zero calibration results.
81
Gain Adjustment
The Internal DMM stores a single new gain correction constant each time this procedure is followed. The gain constant is computed from the calibration value entered for the calibration command and from measurements made automatically during the adjustment procedure. Most measuring functions and ranges have gain adjustment procedures. Only the 100 M range does not have gain calibration procedures. The gain calibration value may be entered through the front panel menu or over the remote interface. See page 53, for an example of how to enter calibration values. Adjustments for each function should be performed ONLY in the order shown in the performance verification table. See Performance Verification Tests earlier in this chapter for the tables used for gain adjustments. Gain Adjustment Considerations The zero adjustment procedure must have been recently performed prior to beginning any gain adjustment procedures. The optional 10 Vdc adjustment should be performed only after servicing the Internal DMMs a-to-d converter or after replacing network A4U101 or calibration RAM A4U505. When performing a 4-wire ohms gain adjustment, a new gain correction constant is also stored for the corresponding 2-wire ohms measurement range. If desired, the 2-wire gain can be adjusted separately after the 4-wire ohms gain calibration is completed. During the ac voltage gain adjustments, some of the dc voltage gain constants are used. Perform the dc voltage gain calibration before the ac voltage gain calibration. Note: Never turn off the instrument during a Gain Adjustment. This may cause calibration memory for the present function to be lost.
82
Valid Gain Adjustment Input Values Gain adjustment can be accomplished using the following input values.
Function DC VOLTS OHMS, OHMS 4W DC CURRENT AC VOLTS [1] AC CURRENT Frequency Range 100 mV to 100 V 300 V 100 to 10 M 10 mA to 1 A 10 mV to 100 V 300 V 1A Any Valid Calibration Input Values 0.9 to 1.1 x Full Scale 250 V to 303 V 0.9 to 1.1 x Full Scale 0.9 to 1.1 x Full Scale 0.9 to 1.1 x Full Scale 95 V to 303 V 9 mA to 11 mA Any Input > 100 mV rms, 1 kHz 100 kHz
1 kHz 10% for the 1 kHz calibration, 45 kHz 100 kHz for the 50 kHz calibration, and 10 Hz 10% for the 10 Hz calibration.
83
Gain Adjustment Procedure Adjustment for each function should be performed only in the order shown in the performance verification table. The performance verification tables used for gain adjustments start on page 77. Review the Test Considerations (page 71) and Gain Adjustment Considerations (page 82) sections before beginning this test. Configuration: 1 2 DC functions 612 digits AC functions LF 3 HZ:SLOW (in the Advanced menu)
Select Channel 210. Configure the channel to each function and range shown in the gain verification tables (pages 75 79). Apply the input signal shown in the Input column of the appropriate verification table. Note: Always complete tests in the same order as shown in the appropriate verification table.
Press to enter the calibration menu. Press begin the adjustment procedure.
again to
4 5 6
Use the knob, and to set the number in the display to the actual input value and press . Perform the appropriate Gain Verification Test to check the calibration results. Repeat steps 1 through 6 for each gain verification test point shown in the tables. Note: Each range in the gain adjustment procedure takes less than 20 seconds to complete.
84
85
Continued from previous page... 1 Reverse the cable connections to the calibrator to create a -10 Vdc voltage standard. You must physically reverse the cables. DO NOT switch the output polarity of the calibrator. Perform a 10V DC gain calibration. Press to enter the calibration menu. Press again to begin the adjustment procedure. Be sure to allow time for thermal offsets to stabilize (usually about 1 minute). Use the knob and to set the number in the display to the difference of the calibrator output and the measured offset (from step 3) and press . Using the previous example values, enter 10 V minus 10.001 volts or 10.000990 volts. When the adjustment finishes, verify that new readings fall within 30 V of the calibrator output minus the offset.
86
Module Reference
87
Relay Verification
There are two methods you can use to verify relays: Read the relay cycle count. Measure the relay contact resistance.
88
(Optional)
Module Reference
Tests 1 - 5:
See the diagram on page 90 for the required connections for each test (be sure to probe the components at the indicated location). For these measurements, the 34901A is not installed in the 34970A. Record the 4-wire ohms measurements from the external DMM in the table below. Note: The connections to the external DMM are different for each of Tests 1, 2, 4, and 5. Be sure to verify the connections shown in the table below for each of the four tests.
External DMM Ohmmeter Connections Test # 1 2 3 4 5 F501 F502 HI L401 L402 LO J101, C14 J101, C15 HI Sense L401 L402 LO Sense J101, C14 J101, C15 Measured Value __________ Ohms __________ Ohms __________ Ohms Ch 21 I Ch 22 I __________ Ohms __________ Ohms
89
Note: Connect bare copper wires (approximately 3 cm in length) to the I terminals of Channels 21 and 22 as shown below. These wires will be used to make shorts across the channels in Tests 6 through 39.
Note: Be sure to probe the components at the indicated locations on the module.
Probe here for F501 measurement. Probe here for F502 measurement.
90
Tests 6 - 8:
Make the connections to the 34901A as shown in the diagram below. Be sure to route your wiring for proper strain relief and install the module cover. Install the 34901A in slot 200 of the 34970A. Open all channels on the module by performing a Factory Reset (press and select Recall State; press again and select Factory Reset). Configure Channel 20 as follows: DC volts, 10 volt range, and 512 digits.
Module Reference
Connections for 34901A Verification Tests 6 through 39 Enable reading monitoring by pressing on the selected channel (or use the ROUTe:MON command). Record the 4-wire ohms measurements from the external DMM in the following table.
External DMM Ohmmeter Connections Test Channel # Configured 6 7 8 Ch 20 Ch 20 HI P2 P2 LO P2 P2 HI Sense P1 P3 LO Sense P1 P3 Measured Value __________ Ohms __________ Ohms __________ Ohms Test Limit 2.00 2.00 Relay Measured K421 K422
91
Tests 9 - 10:
Open all channels on the module by performing a Factory Reset. Configure Channel 10 (module in slot 200) as follows: 4-wire ohms, 1 k range, and 512 digits. Enable reading monitoring by pressing on the selected channel (or use the ROUTe:MON command). Record the 4-wire ohms measurements from the external DMM in the following table.
External DMM Ohmmeter Connections
HI P2
LO P2
HI Sense P3
LO Sense P3
Open all channels on the module by performing a Factory Reset. For each test, close only the channel shown in the Channel Closed column below (module in slot 200). Turn the Monitor Mode off and select Banks Joined from the Advanced menu. Record the 4-wire ohms measurements from the external DMM in the following table.
External DMM Ohmmeter Connections
Test # 11 12 13 14 15 16 17
Channel Closed* Ch 1 Ch 2 Ch 3 Ch 4 Ch 5 Ch 6 Ch 7
HI P2 P2 P2 P2 P2 P2 P2
LO P2 P2 P2 P2 P2 P2 P2
HI Sense P1 P1 P1 P1 P1 P1 P1
LO Sense P1 P1 P1 P1 P1 P1 P1
Measured Value __________ Ohms __________ Ohms __________ Ohms __________ Ohms __________ Ohms __________ Ohms __________ Ohms
* Only the channel currently under test should be closed at one time. All other channels should be open.
Module Reference
External DMM Ohmmeter Connections Test # 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 Ch 22 Channel Closed* Ch 8 Ch 9 Ch 10 Ch 11 Ch 12 Ch 13 Ch 14 Ch 15 Ch 16 Ch 17 Ch 18 Ch 19 Ch 21 HI P2 P2 P2 P2 P2 P2 P2 P2 P2 P2 P2 P2 P4 LO P2 P2 P2 P2 P2 P2 P2 P2 P2 P2 P2 P2 P4 HI Sense P1 P1 P1 P1 P1 P1 P1 P1 P1 P1 P1 P1 P5 LO Sense P1 P1 P1 P1 P1 P1 P1 P1 P1 P1 P1 P1 P5 Measured Value __________ Ohms __________ Ohms __________ Ohms __________ Ohms __________ Ohms __________ Ohms __________ Ohms __________ Ohms __________ Ohms __________ Ohms __________ Ohms __________ Ohms __________ Ohms __________ Ohms P5 __________ Ohms __________ Ohms Test Limit 2.00 2.00 2.00 2.00 2.00 2.00 2.00 2.00 2.00 2.00 2.00 2.00 2.00 2.00 Relay Measured K408 K409 K410 K411 K412 K413 K414 K415 K416 K417 K418 K419 K522 K522
* Only the channel currently under test should be closed at one time. All other channels should be open.
93
Tests 34 - 36:
Close only channels Channels 20 and 22. Remove the 34901A from the 34970A and do not reinstall it for these tests. On connector J101, remove the jumper between pins C14 and C15 (the top jumper shown in the diagram on page 91). On the remaining jumper connected to J101 (the bottom jumper shown in the diagram), move the end of the jumper from pin C12 to pin C16; the jumper should now short pins C13 and C16 together. Cut, but do not remove, the copper shorts on Channels 21 and 22 (the wires will be used for the 4-wire ohms measurements below). Add a copper short between the L and H terminals on Channel 20. Record the measured value as Test 34 in the table below. Using the external DMM, make a 4-wire ohms measurement between the L and I terminals on Channel 21. Record the measured value as Test 35 in the table below.
External DMM Ohmmeter Connections Test # 34 35 36 Channel Closed Ch 20* Ch 22* HI P3 Ch 21 I LO P3 Ch 21 L HI Sense P2 Ch 21 I LO Sense P2 Ch 21 L Measured Value __________ Ohms __________ Ohms __________ Ohms Test Limit 2.00 2.00 Relay Measured K420 K523
* The latching relays remain closed when the module is removed from the 34970A.
94
Test 37:
Install the 34901A in slot 200 of the 34970A. Select and configure Channel 21 as follows: DC current, 1 amp range, and 512 digits. Enable reading monitoring by pressing on the selected channel (or use the ROUTe:MON command). Record the 4-wire ohms measurement from the external DMM in the following table.
Module Reference
External DMM Ohmmeter Connections Test Channel # Configured 37 Ch 21 HI P4 LO P4 HI Sense P5 LO Sense P5 Measured Value __________ Ohms Test Limit 2.00 Relay Measured K521
Tests 38 - 39:
Open all channels on the module by performing a Factory Reset. Close Channel 21 (module in slot 200). Remove the 34901A from the 34970A and do not reinstall it for the remaining tests. Using the external DMM, make a 4-wire ohms measurement between the L and I terminals on Channel 22. Record the measured value as Test 38 in the following table.
External DMM Ohmmeter Connections Test # 38 39 Channel Closed Ch 21 HI Ch 22 I LO Ch 22 L HI Sense Ch 22 I LO Sense Ch 22 L Measured Value __________ Ohms __________ Ohms Test Limit 2.00 Relay Measured K524
95
(Optional)
Tests 1 - 4:
See the diagram on page 97 for the required connections for each test (be sure to probe the components at the indicated location). For these measurements, the 34902A is not installed in the 34970A. Record the 4-wire ohms measurements from the external DMM in the table below. Note: The connections to the external DMM are different for each of Tests 1, 2, and 3. Be sure to verify the connections shown in the table below for each of the three tests.
External DMM Ohmmeter Connections Test # 1 2 3 4 HI L300 L301 L302 LO J101, C12 J101, C14 J101, C15 HI Sense L300 L301 L302 LO Sense J101, C12 J101, C14 J101, C15 Measured Value __________ Ohms __________ Ohms __________ Ohms __________ Ohms
96
Module Reference
97
Tests 5 - 8:
Make the connections to the 34902A as shown in the diagram below. Be sure to route your wiring for proper strain relief and install the module cover. Install the 34902A in slot 200 of the 34970A. Open all channels on the module by performing a Factory Reset (press and select Recall State; press again and select Factory Reset). Configure Channel 16 as follows: DC volts, 10 volt range, and 512 digits.
Connections for 34902A Verification Tests 5 through 27 Enable reading monitoring by pressing on the selected channel (or use the ROUTe:MON command). Record the 4-wire ohms measurements from the external DMM in the following table.
External DMM Ohmmeter Connections Test Channel # Configured 5 6 7 8 Ch 16 Ch 16 HI P3 LO P3 HI Sense P1 LO Sense P1 Measured Value __________ Ohms __________ Ohms P2 __________ Ohms __________ Ohms Test Limit 2.00 2.00 Relay Measured K326 K327
98
Tests 9 - 10:
Open all channels on the module by performing a Factory Reset. Configure Channel 08 (module in slot 200) as follows: 4-wire ohms, 1 k range, and 512 digits. Enable reading monitoring by pressing on the selected channel (or use the ROUTe:MON command). Record the 4-wire ohms measurements from the external DMM in the following table.
Module Reference
External DMM Ohmmeter Connections Test Channel # Configured 9 10 Ch 08 HI P3 LO P3 HI Sense P2 LO Sense P2 Measured Value __________ Ohms __________ Ohms Test Limit 2.00 Relay Measured K328
4
Tests 11 - 27: Open all channels on the module by performing a Factory Reset. For each test, close only the channels shown in the Channels Closed column below (module in slot 200). Turn the Monitor Mode off and select Banks Joined from the Advanced menu. Record the 4-wire ohms measurements from the external DMM in the following table.
External DMM Ohmmeter Connections Test # 11 12 13 14 15 Channels Closed* Ch 16 & 1 Ch 16 & 2 Ch 16 & 3 Ch 16 & 4 Ch 16 & 5 HI P3 P3 P3 P3 P3 LO P3 P3 P3 P3 P3 HI Sense P1 P1 P1 P1 P1 LO Sense P1 P1 P1 P1 P1 Measured Value __________ Ohms __________ Ohms __________ Ohms __________ Ohms __________ Ohms Test Limit 2.00 2.00 2.00 2.00 2.00 Relay Measured K301 K302 K303 K304 K305
* Only the channels currently under test should be closed at one time. All other channels should be open.
99
External DMM Ohmmeter Connections Test # 16 17 18 19 20 21 22 23 24 25 26 27 Channels Closed* Ch 16 & 6 Ch 16 & 7 Ch 16 & 8 Ch 16 & 9 Ch 16 & 10 Ch 16 & 11 Ch 16 & 12 Ch 16 & 13 Ch 16 & 14 Ch 16 & 15 Ch 16 & 15 HI P3 P3 P3 P3 P3 P3 P3 P3 P3 P3 P3 LO P3 P3 P3 P3 P3 P3 P3 P3 P3 P3 P3 HI Sense P1 P1 P1 P1 P1 P1 P1 P1 P1 P1 P4 LO Sense P1 P1 P1 P1 P1 P1 P1 P1 P1 P1 P4 Measured Value __________ Ohms __________ Ohms __________ Ohms __________ Ohms __________ Ohms __________ Ohms __________ Ohms __________ Ohms __________ Ohms __________ Ohms __________ Ohms __________ Ohms Test Limit 2.00 2.00 2.00 2.00 2.00 2.00 2.00 2.00 2.00 2.00 2.00 2.00 Relay Measured K306 K307 K308 K309 K310 K311 K312 K313 K314 K315 K316
* Only the channels currently under test should be closed at one time. All other channels should be open.
100
(Optional)
1 2 3 4 5
Module Reference
Be sure to read Plug-in Module Test Considerations on page 87. Install the 34903A module in slot 100. Close Channels 01 through 20. Remove the module from the 34970A. Measure the resistance from the CM terminal to the NO terminal on each channel. Install the module in slot 100. Open Channel 01 through 20. Remove the module from the 34970A. Measure the resistance from the CM terminal to the NC terminal on each channel.
Note: In general, a new relay should have a contact resistance of less than 0.2. Relays with contact resistance in excess of 1.2 should be replaced.
101
(Optional)
102
Tests 1 - 32:
Open all channels on the module by performing a Factory Reset (press and select Recall State; press again and select Factory Reset). For each test, close only the channels shown in the Channels Closed column below (module in slot 200). Record the 4-wire ohms measurements from the external DMM in the following table. Note: To prepare the module between tests, press and hold down until Card Reset is displayed on the front panel
Module Reference
External DMM Ohmmeter Connections Test # 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 Channels Closed* Ch 11 & 41 Ch 12 & 42 Ch 13 & 43 Ch 14 & 44 Ch 15 & 45 Ch 16 & 46 Ch 17 & 47 Ch 18 & 48 Ch 21 & 41 Ch 22 & 42 Ch 23 & 43 Ch 24 & 44 Ch 25 & 45 Ch 26 & 46 Ch 27 & 47 Ch 28 & 48 HI P1 P1 P1 P1 P1 P1 P1 P1 P2 P2 P2 P2 P2 P2 P2 P2 LO P1 P1 P1 P1 P1 P1 P1 P1 P2 P2 P2 P2 P2 P2 P2 P2 HI Sense P4 P4 P4 P4 P4 P4 P4 P4 P4 P4 P4 P4 P4 P4 P4 P4 LO Sense P4 P4 P4 P4 P4 P4 P4 P4 P4 P4 P4 P4 P4 P4 P4 P4 Measured Value __________ Ohms __________ Ohms __________ Ohms __________ Ohms __________ Ohms __________ Ohms __________ Ohms __________ Ohms __________ Ohms __________ Ohms __________ Ohms __________ Ohms __________ Ohms __________ Ohms __________ Ohms __________ Ohms Test Limit 2.00 2.00 2.00 2.00 2.00 2.00 2.00 2.00 2.00 2.00 2.00 2.00 2.00 2.00 2.00 2.00 Relay Measured K101 K102 K103 K104 K105 K106 K107 K108 K201 K202 K203 K204 K205 K206 K207 K208
* Only the channels currently under test should be closed at one time. All other channels should be open.
External DMM Ohmmeter Connections Test # 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 Channels Closed* Ch 31 & 41 Ch 32 & 42 Ch 33 & 43 Ch 34 & 44 Ch 35 & 45 Ch 36 & 46 Ch 37 & 47 Ch 38 & 48 Ch 41 & 11 Ch 42 & 12 Ch 43 & 13 Ch 44 & 14 Ch 45 & 15 Ch 46 & 16 Ch 47 & 17 Ch 48 & 18 HI P3 P3 P3 P3 P3 P3 P3 P3 P4 P4 P4 P4 P4 P4 P4 P4 LO P3 P3 P3 P3 P3 P3 P3 P3 P4 P4 P4 P4 P4 P4 P4 P4 HI Sense P4 P4 P4 P4 P4 P4 P4 P4 P1 P1 P1 P1 P1 P1 P1 P1 LO Sense P4 P4 P4 P4 P4 P4 P4 P4 P1 P1 P1 P1 P1 P1 P1 P1 Measured Value __________ Ohms __________ Ohms __________ Ohms __________ Ohms __________ Ohms __________ Ohms __________ Ohms __________ Ohms __________ Ohms __________ Ohms __________ Ohms __________ Ohms __________ Ohms __________ Ohms __________ Ohms __________ Ohms Test Limit 2.00 2.00 2.00 2.00 2.00 2.00 2.00 2.00 2.00 2.00 2.00 2.00 2.00 2.00 2.00 2.00 Relay Measured K301 K302 K303 K304 K305 K306 K307 K308 K401 K402 K403 K404 K405 K406 K407 K408
* Only the channels currently under test should be closed at one time. All other channels should be open.
104
(Optional)
Module Reference
3 4 5 6 7
105
(Optional)
Test 1:
See the diagram on page 107 for the required connections for this test (be sure to probe the inductor at the indicated location). For this measurement, the 34908A is not installed in the 34970A. Record the 4-wire ohms measurements from the external DMM in the table below.
External DMM Ohmmeter Connections Test # 1 HI L400 LO J101, C12 HI Sense L400 LO Sense J101, C12 Measured Value __________ Ohms
106
Module Reference
107
Tests 2 - 3:
Make the connections to the 34908A as shown in the diagram below. Be sure to route your wiring for proper strain relief and install the module cover. Install the 34908A in slot 200 of the 34970A. Open all channels on the module by performing a Factory Reset (press and select Recall State; press again and select Factory Reset). Configure Channel 01 as follows: DC volts, 10 volt range, and 512 digits.
Connections for 34908A Verification Tests 2 and 3 Enable reading monitoring by pressing on the selected channel (or use the ROUTe:MON command). Record the 4-wire ohms measurements from the external DMM in the following table.
External DMM Ohmmeter Connections Test Channel # Configured 2 3 Ch 01 HI H Com LO L Com HI Sense H Com LO Sense L Com Measured Value __________ Ohms __________ Ohms Test Limit 2.00 Relay Measured K421
108
Tests 4 - 43:
Make the connections to the 34908A as shown in the diagram below. Connect copper shorts between all channels as shown. Be sure to route your wiring for proper strain relief and install the module cover. Install the 34908A in slot 200 of the 34970A.
Module Reference
Connections for 34908A Verification Tests 4 through 43 For each test shown in the table starting on the next page, close only the channels shown in the Channels Closed column (closing one channel will open the previously closed channel). Record the 4-wire ohms measurements from the external DMM in the table.
109
External DMM Ohmmeter Connections Test Channels # Closed* 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 Ch 1 Ch 2 Ch 3 Ch 4 Ch 5 Ch 6 Ch 7 Ch 8 Ch 9 Ch 10 Ch 11 Ch 12 Ch 13 Ch 14 Ch 15 Ch 16 Ch 17 Ch 18 Ch 19 Ch 20 HI H Com H Com H Com H Com H Com H Com H Com H Com H Com H Com H Com H Com H Com H Com H Com H Com H Com H Com H Com H Com LO L Com L Com L Com L Com L Com L Com L Com L Com L Com L Com L Com L Com L Com L Com L Com L Com L Com L Com L Com L Com HI Sense H Com H Com H Com H Com H Com H Com H Com H Com H Com H Com H Com H Com H Com H Com H Com H Com H Com H Com H Com H Com LO Sense L Com L Com L Com L Com L Com L Com L Com L Com L Com L Com L Com L Com L Com L Com L Com L Com L Com L Com L Com L Com Measured Value __________ Ohms __________ Ohms __________ Ohms __________ Ohms __________ Ohms __________ Ohms __________ Ohms __________ Ohms __________ Ohms __________ Ohms __________ Ohms __________ Ohms __________ Ohms __________ Ohms __________ Ohms __________ Ohms __________ Ohms __________ Ohms __________ Ohms __________ Ohms Test Limit 2.00 2.00 2.00 2.00 2.00 2.00 2.00 2.00 2.00 2.00 2.00 2.00 2.00 2.00 2.00 2.00 2.00 2.00 2.00 2.00 Relay Measured K401, K422 K402, K422 K403, K422 K404, K422 K405, K422 K406, K422 K407, K422 K408, K422 K409, K422 K410, K422 K411, K422 K412, K422 K413, K422 K414, K422 K415, K422 K416, K422 K417, K422 K418, K422 K419, K422 K420, K422
* Only the channels currently under test should be closed at one time. All other channels should be open.
Module Reference
External DMM Ohmmeter Connections Test Channels # Closed* 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 Ch 21 Ch 22 Ch 23 Ch 24 Ch 25 Ch 26 Ch 27 Ch 28 Ch 29 Ch 30 Ch 31 Ch 32 Ch 33 Ch 34 Ch 35 Ch 36 Ch 37 Ch 38 Ch 39 Ch 40 HI H Com H Com H Com H Com H Com H Com H Com H Com H Com H Com H Com H Com H Com H Com H Com H Com H Com H Com H Com H Com LO L Com L Com L Com L Com L Com L Com L Com L Com L Com L Com L Com L Com L Com L Com L Com L Com L Com L Com L Com L Com HI Sense H Com H Com H Com H Com H Com H Com H Com H Com H Com H Com H Com H Com H Com H Com H Com H Com H Com H Com H Com H Com LO Sense L Com L Com L Com L Com L Com L Com L Com L Com L Com L Com L Com L Com L Com L Com L Com L Com L Com L Com L Com L Com Measured Value __________ Ohms __________ Ohms __________ Ohms __________ Ohms __________ Ohms __________ Ohms __________ Ohms __________ Ohms __________ Ohms __________ Ohms __________ Ohms __________ Ohms __________ Ohms __________ Ohms __________ Ohms __________ Ohms __________ Ohms __________ Ohms __________ Ohms __________ Ohms Test Limit 2.00 2.00 2.00 2.00 2.00 2.00 2.00 2.00 2.00 2.00 2.00 2.00 2.00 2.00 2.00 2.00 2.00 2.00 2.00 2.00 Relay Measured K401, K422 K402, K422 K403, K422 K404, K422 K405, K422 K406, K422 K407, K422 K408, K422 K409, K422 K410, K422 K411, K422 K412, K422 K413, K422 K414, K422 K415, K422 K416, K422 K417, K422 K418, K422 K419, K422 K420, K422
* Only the channels currently under test should be closed at one time. All other channels should be open.
Note: If the first 20 or last 20 relays have high resistance values, it is likely that relay K422 is bad.
111
3 4 5
112
34901A
Module Reference
2 3
Install the plug-in module in the mainframe in slot 200. Apply power and allow a 2 hour warm-up. Set 10 k thermistor measurements on Channels 206 and 217 (or 206 and 211). Before executing each test, you must press to enable reading monitoring on the selected channel (or use the ROUTe:MON command from the remote interface). Press to enter the calibration menu. Press begin the adjustment procedure. Verify the adjustment (see page 112). again to
4 5
113
34907A
To DMM
For each analog output, set each output value in the table below. Compare measurement results to the appropriate test limits shown in the table.
DAC Output 0.000 V 10.000 V Quick Check Q Measured Output 0.000 V 10.000 V Error from Nominal 1 year 20 mV 45 mV
114
Module Reference
5 6 7
Use the knob and to set the number in the display to the measured output value and press to complete the adjustment. Set the analog output to 10.000 V. Press to enter the calibration menu. Press begin the adjustment procedure. again to
8 9
Use the knob, and to set the number in the display to the measured output value and press to complete the adjustment. Repeat steps 1 through 8 for Channel 105.
10 Perform the Voltage Output Verification Test on page 114 to verify the adjustment.
115
116
Theory of Operation
Theory of Operation
This chapter is organized to provide descriptions of the circuitry contained on each schematic shown in chapter 8. A block diagram overview is provided followed by more detailed descriptions of the circuitry contained in the schematics chapter. System Block Diagram, on page 119 Floating Logic, on page 120 Memory, on page 123 Earth-Referenced Logic, on page 124 Power Supplies, on page 125 Front Panel, on page 127 Backplane, on page 128 Internal DMM, on page 129 Switch Modules, on page 138 Multifunction Module, on page 151 The self-test procedures are described in chapter 6.
118
119
Floating Logic
Unless otherwise noted, components in this discussion are located on the A1 circuit assembly (34970-66501). The schematics are included in Chapter 8 starting on page 224. The floating common logic controls the operation of the entire instrument. All measurement control and remote interface command interpretation is performed in the main controller, U205. The front panel controller, the I/O controller, and all of the plug-in module controllers, act as slaves to U205. The floating common logic is comprised of the main controller U205, custom ASIC U209, calibration memory U201, 12 MHz clock oscillator U204, and microprocessor supervisor U104. The microprocessor supervisor U104 performs the following functions: Monitors the regulated 5V floating supply and generates the reset signal for main controller U205 when the voltage drops below operating levels. Monitors the unregulated side of the 5V floating supply in order to generate an early warning signal (PWRFAIL) when power is lost. Provides automatic switch over to the battery BT101 for the +5V_NV supply when the 5V_FLT supply drops below the battery voltage. Blocks the main controllers write signal (WR_N) while the 5V_FLT supply is below operating level.
120
The main controller, U205, is a 16-bit microcontroller incorporating many built-in features: A 10-bit, successive approximation ADC with selectable inputs is used to convert two signals: FLASH and FRQRNG. The FLASH signal is the residual charge on the main integrating ADC output from the internal DMM assembly (A4). The FRQRNG signal (also from the internal DMM) is used to make voltage ranging decisions for frequency and period measurements. A pulse-width-modulation port, after filtering the 23 kHz output with R221,C224, and R259, produces DC voltages between 0V and 5V. This voltage, PREADJ, is used to adjust the precharge amplifier offset voltage in U101 on the internal DMM assembly (A4). A full, UART controlled, serial port is used to communicate with the I/O processor through optoisolators U303 and U214. Data is sent in an 11-bit frame at a rate of 187.5 kbits/second. The 11-bit frame is configured for one start bit, nine data bits, and one stop bit. A timer is used to measure the power line frequency on LSENSE. Frequencies between 55 Hz and 66 Hz result in the use of a 60 Hz standard for the DMM integration period. All other frequencies will result in the use of a 50 Hz standard. A 16-bit counter counts pulses on CNT to create, along with the 8-bit counter in U209, a 24-bit counter for the internal DMM.
121
The custom ASIC, U209, provides: Memory Address mapping The main controller multiplexes address and data on the same bus. U209 latches the address and drives a separate memory address bus (MA(19:1)). U209 allows the main controller to access a much larger memory space than its 16-bit address bus would allow. It also partitions memory into separate data and instruction segments and a mappable segment that can be used for either data or instructions. Communications U209 provides three serial communication ports. A 187.5 kbit/second, 9 data bit, UART for communicating with the isolated backplane (FLT_BPDO and FLT_BPDI_N). A duplicate, 9-bit UART to communicate with the front panel (FPDO and FPDI). And a simple, 1.5 Mbit/second, clocked shift-register to control the configuration registers on the DMM assembly (CFG_SCK, CFG_DO, and CFG_DI). Real Time Clock U209 provides time of day and date, a periodic interrupt, and a squarewave generator. The date is based on a 100 year calendar (it accurately tracks leap years until its two digit year counter rolls over). U209 uses the battery-backed +5V_NV supply. DMM support U209 provides conversion logic for the analog-to-digital converter and a counter for both the ADC and the frequency measurement features of the DMM. When used for the ADC, the COMP input functions both as a clocked comparator and the slope counter input. When used for frequency measurement, FREQIN is the input to the counter. In both cases, the 8-bit counter produces the lower bits of a 24-bit counter. The counter overflow signal, CNT, is counted by U205 to produce the upper 16-bits of the count. The SYNC signal produced by U205, is used to latch the count. Device U201 is a 512 byte, ferroelectric RAM that is accessed via a serial interface. This device only contains data relevant to the A1 assembly. This data is combined with data retrieved from A4U450 to completely calibrate the internal DMM. The two devices, although on different assemblies, share the same I/O signals, CALSCK and CALDAT, that allow them to be read (and written) by U205.
122
Memory
Unless otherwise noted, components in this discussion are located on the A1 circuit assembly (34970-66501). The schematics are included in Chapter 8 starting on page 224. The main controller, U205, uses 512 Kbytes of ROM and 544 Kbytes of RAM. ASIC U209 provides the memory mapping that allows access to this large memory space. The memory map as seen by the main controller is as follows:
Address 0100H thru 1FBFH 0100H thru 1FBFH 1FC0H thru 1FDFH 2000H thru 5FFFH 6000H thru 7FFFH 6000H thru 7FFFH 8000H thru FFFFH Maps to 00100H thru 01FBFH in RAM for data fetches 00100H thru 01FBFH in ROM for instr. fetches registers in U209 02000H thru 05FFFH in ROM 06000H thru 07FFFH in RAM for data fetches 06000H thru 07FFFH in ROM for instr. fetches any selected 32 Kbyte page of ROM or RAM
The memory is organized with a 16-bit data bus (AD(15:0)) and a 19-bit address bus (MA(19:1)). The memory address is produced by U209 latching the address present on AD(15:0) when U205 asserts ALE_FLT. Reads of memory are always 16-bits wide (there is a single RD_N signal). Writes, however, can be byte-wide and therefore U209 produces both a high-byte write strobe, MWRH_N, and a low-byte write strobe, MWRL_N. These write signals are based on the latched 0-bit of the address and the BHE_N signal produced by the main controller, U205. The ROM memory consists of a single, 256Kx16 device, U401. The RAM memory consists of five devices: U402, U403, U404, U405, and U410. A uniform, 256Kx16, memory block is formed by the four 128Kx8 devices, U402 through U405. A separate, 32Kx8 block is formed by U410 and is available through special programming of U209.
123
Earth-Referenced Logic
Unless otherwise noted, components in this discussion are located on the A1 circuit assembly (34970-66501). The schematics are included in Chapter 8 starting on page 224. The earth-referenced logic circuits provide all rear panel input/output capability. Microprocessor U305 handles GPIB (IEEE-488) control through bus interface chip U309 and bus receiver/driver chips U310 and U311. The RS-232 interface is controlled by U305 through U307. RS-232 transceiver chip U308 provides the required level shifting to approximate 9 volt logic levels through on-chip charge-pump power supplies using capacitors C317 through C320. Communication between the earth referenced logic interface circuits and the floating measurement logic is accomplished through an optically-isolated bi-directional serial interface. Isolator U214 couples data from U305 to microprocessor U205. Isolator U303 couples data from U305 to microprocessor U205. U305 also: Controls power to the backplane as well as all rear panel interfaces (GPIB, RS-232, Alarms, External Triggers). Backplane power is turned on or off based upon commands received from U205. Drives the backplane reset signal (BPRST) based upon commands from U205. Monitors the backplane service request (BPSRQ_N) and reports to U205 when it is asserted. U306B, U306C, U306D, and U306E drive the alarm outputs. The alarm is a low true signal at the sub miniature D connector on the rear panel. U306F drives the channel closed output signal. The external trigger input is buffered by U304C and U304D.
124
Power Supplies
Unless otherwise noted, components in this discussion are located on the A1 circuit assembly (34970-66501). The schematics are included in Chapter 8 starting on page 224. The instrument uses two types of power supplies: floating supplies and earth referenced supplies. The floating supply outputs are 18 Vdc, +5 Vdc, and a 6 Vrms center tapped filament supply for the vacuum fluorescent display. The earth referenced and backplane circuits are powered from a single +5 Vdc supply. The ac mains are connected by module P1. This module includes the functions of mains connection and line voltage selection (100/120/ 220/240). The internal DMM automatically configures for the applied line frequency by counting the frequency of the output of clamp circuit CR106, R102, C103 (LSENSE). The +5 volt floating supply is produced by bridge rectifier CR105, filter capacitor C104, and regulator U103. The output of CR105 is sensed by U104 and compared to the the voltage from battery BT101. U104 turns on the +5 V floating supply through Q120. If the output of CR105 falls below 6.8 V, U104 provides a PWRFAIL signal to the main processor. At initial power on, U104 resets the main processor with the FLT_RST line. This supply powers all floating logic. The internal DMM relay drive circuits are also powered from this supply. The floating 18 volt supplies are produced by bridge rectifier CR109, filter capacitors C107 and C109, and regulators U105 and U106. These supplies are used to power all measuring circuits. In addition, the vacuum fluorescent display is driven from the 18 volt supplies. A separate winding of T1 provides a center tapped 6 Vrms filament supply for the display. Q110A and Q110B turn on and off the filament supply in response to the FILPWR signal from the main controller through U107A.
125
The 5 volt earth referenced supply (+5V_ER) is produced by rectifier CR101, CR102, CR103, CR104, and regulator U101. This supply is earth referenced by the screw which mounts the PC board to the instrument chassis. The unswitched +5V_ER supplies U305, U303, U320, and U302. The rear-panel interfaces (GPIB, RS-232, Alarms, and external triggers) are powered from the switched +5V_BP supply to ensure that when power is turned off no voltages are present at the interfaces. The +5V_ER supply is switched by Q101A and Q101B to create the +5V_BP (backplane) and fan power supplies. Undervoltage sensor U102 provides the earth reference controller reset at initial power on.
126
Front Panel
Unless otherwise noted, components in this discussion are located on the A2 circuit assembly (34970-66502). The schematics are included in Chapter 8 starting on page 229. The front-panel circuits consist of vacuum fluorescent display control, display high voltage drivers, and keyboard scanning. Communication between the front panel and floating logic circuits is accomplished through a 2-wire bi-directional serial interface. The front-panel logic operates from -13 volts (logic 1) and -18 volts (logic 0). The two serial communication signals are level shifted by comparator U6 from the floating logic 0 V to 5 V levels to the -18 V to -13 V levels present on the front panel assembly. The front panel logic high supply (-13 volts) is produced from the -18 volt supply by voltage regulator U7. Display anode and grid voltages are +18 volts for an on segment and -18 volts for an off segment. The -11 V cathode bias for the display is provided by the main pc boards filament winding center tap bias circuit A1CR108, A1R106, and A1C106 shown on the power supply schematic (see page 224). Keyboard scanning is accomplished through a conventional scanned row-column key matrix. Keys are scanned by outputting data at microprocessor U1 port pins P0.0 through P0.4 to poll each key column for a key press. Column read-back data are read by the microprocessor at port pins P2.0 through P2.3 for decoding and communication to the floating logic circuits. Rotary knob quadrature inputs are read directly by the microprocessor port pins P2.6 and P2.7. The standby power switch, S19, provides a low true signal to main controller A1U205. In turn, A1U205 takes actions that either place the instrument in the standby mode or on mode. In standby, both the filament supply to the front panel and the +5V_BP supply to the backplane, rear panel interfaces, and fan are turned off.
127
Backplane
Unless otherwise noted, components in this discussion are located on the A3 circuit assembly (34970-66503). The schematics are included in Chapter 8 starting on page 231. The backplane contains three connectors, P101, P102, and P103 for connection to the plug-in modules. The parallel lines in these connectors are divided into two groups to form the analog bus and digital bus.
Analog Bus
The analog bus connects the signals from the plug-in modules to the Internal DMM. There are five lines in the analog bus, HI, LO, OHMS_HI, OHMS_LO, and AMPS. The HI and LO lines are protected from overvoltages by E101, E102, RV101, RV102, R101, R160, L101, L102, and C109. P105 makes the analog bus connection to the internal DMM.
Digital Bus
The digital bus uses 10 lines for communication and control. P104 makes the digital bus connection to the earth referenced logic and floating logic.
P101, P102, P103 Pins A6, A7, A8 B6 C6 A5, B5, B7, C5, C7 A3, B3, C3 A1, B1, C1 B8 C8
Comments Unique binary code for each slot. Serial module data from the floating logic. Serial module data to the floating logic. Earth referenced digital ground. Earth referenced module power supply. Earth referenced zap return ground. Module reset from the earth referenced logic. Module service request to the earth referenced logic.
128
Internal DMM
DMM Block Diagram
The internal DMM block diagram is shown on the system block diagram on page 221. A portion of the block diagram is shown below.
5
Backplane
Main Controller
129
Input
Unless otherwise noted, components in this discussion are located on the A4 circuit assembly (34970-66504). The schematics are included in Chapter 8 starting on page 234. The purpose of the Input section is to connect the Input HI terminal to the various measuring functions. This is accomplished through K102, K103, and K104. Additionally, connections are made for the 4-wire ohms HI Sense and LO Sense inputs. Shunt selection (ranging) and voltage sensing are also performed for the current function. The table below shows the state of each relay for each measuring function. All relay coils are driven from U150.
Function 0.1 V 10 Vdc 100 V 300 Vdc 2-Wire Ohms 4-Wire Ohms AC Voltage Frequency/Period 3 A, 1A DC I 100 mA, 10 mA DC I 1A AC I K102 Set Set Reset Reset Set Set Reset Set Reset K103 Set Reset Set Set Reset Reset Reset Reset Reset K104 Set [1] Set Reset [2] Reset [2] Reset Reset Reset Reset Set Sense at: U101-5 U102-12 U101-5 [3] AC_IN AC_IN U101-10 U101-10 AC_IN
[1] K104 will be reset when input resistance is selected to >10,000 M through the menu. [2] K104 will be set for the 100 M range. [3] Configurations shown are for the current source output (HI) terminal. The measurement sense is accomplished through the Sense HI / Sense LO terminals.
130
Input Amplifier
Unless otherwise noted, components in this discussion are located on the A4 circuit assembly (34970-66504). The schematics are included in Chapter 8 starting on page 234. The DC Amplifier circuit is used by every measuring function except frequency and period. Analog switch U101B selects various input signals for measurement by the ADC. Switch U101B has three sources which can be dynamically selected: measure customer input (MC), measure zero input (MZ), and precharge (PRE). The MC state is the actual input measurement. The MZ state measures internal offset voltages which are also present in the MC measurement. The final measurement result is computed from MCMZ. The PRE state is used to precharge internal capacitances to reduce charge injection to the input terminal from the dynamic switching of MC and MZ. Autozero off disables the dynamic switching of the amplifier input. However, a new MZ value is automatically taken whenever a new function or range is selected, even if autozero is turned off. In the dc voltage function, ranging is accomplished through both input relay switching (K102K104) and solid state switching (U101). As a result, the input to the ADC has the same nominal 10 V value for a full scale input on each range. The dc input amplifier is comprised of source follower dual FET U104, amplifier U106, and associated bias circuitry. The feedback resistors U102C and switches U101C select non-inverting amplifier gains of x1, x10, and x100 for the dc input amplifier circuit. Amplifier output ADIN drives the dc input to the a-to-d converter for all measuring functions.
DCV Range 100 mV 1V 10 V 100 V 1000 V U102A Divider U101 Input Pin 5 Pin 5 Pin 5 Pin 8 Pin 8 Amplifier Gain x100 x10 x1 x10 x1 ADC Input 10 V 10 V 10 V 10 V 10 V
1/100 1/100
131
In the DC current function, a current is applied between the Input I and LO terminals. Ranging is accomplished by relay K102 and amplifier gain switching in U101. Since a known resistor (the shunt resister) is connected between these terminals, a voltage proportional to the unknown current is generated. The voltage sensed at R121 is measured by the multimeters dc circuitry. The table below illustrates the dc current measuring function configurations.
DCI Range 1A 100 mA 10 mA Shunt Resistor 0.1 5.1 5.1 U10110 Input 100 mV 510 mV 51 mV Amplifier Gain x100 x10 x100 ADC Input 10V 5.1V 5.1V
Resistance measurements are made by applying a known current through an unknown resistance. The resulting voltage drop across the unknown resistance is then measured by the multimeters dc circuitry. The 100 M range is measured using the known internal 10 M resistance (U102A) in parallel with the unknown input resistance while applying the 500 nA current source. The result is computed from the measured data. The internal 10 M resistance is determined whenever a zero calibration is performed. In the 2-wire ohms function, the voltage drop is measured across the Input HI and Input LO terminals. In the 4-wire ohms function, the voltage is measured across the HI Sense and LO Sense terminals. Lead resistances in series with the current source (Input HILO) are not part of the final measurement. However, they do reduce the available current source compliance voltage for the resistor under test. The ohms current source will become non-linear when the compliance voltage limit is exceeded. The full scale voltage developed across the unknown resistor and the dc amplifier gain for each resistance range are tabulated below.
Ohms Range Voltage Across R 100 100 mV 1V 1 k to 100 k 5V 1 M 5V 10 M 4.5 V 100 M Amplifier Gain x100 x10 x1 x1 x1 ADC Input 10 V 10 V 5V 5V 4.5 V
132
current value is generated by forcing a stable, precise voltage across a stable resistance. The value of the current becomes part of the range gain constant stored during calibration. The +7 V reference voltage is used to generate a stable reference current with U201A. R201 and R202 are the resistance references for the current sources as shown in the table below. The IREF current is used to produce a precise voltage drop across the 28.57 k resistor in U102D-4. The IREF generated using R202 produces an approximate 5 V drop across the 28.57 k resistor. The IREF generated using R201 produces an approximate 0.5 V drop. This voltage is used to force a reference voltage across the selected current source range resistor (5 k, 50 k, 500 k, 1 M) by U201B. The resulting precision current flows through JFET Q202 and protection circuit Q203 to Q211, and CR202 to relay K102 where it is switched to the Input HI terminal for ohms measurements. The protection circuits are designed to protect the ohms current source from inadvertently applied voltages in excess of 1000 V. Protection from large positive voltages is provided by the reverse breakdown voltage of CR202. Protection from large negative voltages is provided by the sum of the collector to base breakdown voltages of Q203, Q205, Q207, and Q209. Bias for these transistors is provided by Q211 and R203 to R206 while negative over voltages are applied.
Ohms Range 100 1 k 10 k 100 k 1 M 10 M 100 M [1] Current 1 mA 1 mA 100 A 10 A 5 A 500 nA 500 nA [1] Open Circuit Compliance Reference Voltage Limit R202 2.5 V 9V R202 2.5 V 9V R202 4V 9V R202 4V 9V R202 8V 9V R201 10 V 14 V R201 5V Isource R U102D 5 k 5 k 50 k 500 k 1 M 1 M 1 M
133
AC Circuit
Unless otherwise noted, components in this discussion are located on the A4 circuit assembly (34970-66504). The schematics are included in Chapter 8 starting on page 234. The multimeter uses a true RMS ac-to-dc converter to measure ac voltages and currents. The ac-to-dc converter changes the input ac voltage to a dc voltage. All voltage ranging is performed in the ac circuit so that the input to the multimeters dc circuitry (AC_OUT) is nominally 2 Vdc for a full scale ac input. The dc amplifier is always configured for x1 gain in ac functions (voltage, current, frequency, and period). Relay K104 connects the ac circuit to either the Input HI terminal or to R121, the current function voltage sense point. Note that the input to the ac circuit may contain a dc bias from the applied ac signal. Input coupling capacitor C301 blocks the dc portion of the input signal. Only the ac component of the input signal is measured by the multimeter. The ac circuit voltage ranging comprises two gain stages U301 and U305/U312. The voltage gains for each stage are tabulated below.
Function ACV, Freq, or Period Range 100 mV 1V 10 V 100 V 300 V 10 mA 100 mA 1A Shunt Resistor 1st Stage x0.2 x0.2 x0.2 x0.002 x0.002 x0.2 x0.2 x0.2 2nd Stage x100 x10 x1 x10 x1 x100 x100 x100 ADC Input 2 Vdc 2 Vdc 2 Vdc 2 Vdc 1.4 Vdc 0.2 Vdc 1 Vdc 2 Vdc
ACI
The 1st stage is a compensated attenuator implementing a gain of x0.2 or x0.002 as selected by U304A and U304D. Each voltage range has a unique 50 kHz frequency response correction produced by a programmable variable capacitor connected across R304.
134
The programmable capacitance is implemented by varying the signal level across a compensating capacitor. In the x0.2 configuration, low frequency gain is set by R301, R302, and R304. The variable gain element U302/U303 essentially varies the value of C306 from 0 to 1 times its value in 256 steps. The exact gain constant is determined during the 50 kHz ac voltage range calibration procedure. In the x0.002 configuration, low frequency gain is set by R301, R302, and R303. The variable gain element U302/U303 essentially varies the value of C305 plus C306 from 0 to 1 times their value in 256 steps. The exact gain constant is determined during the 50 kHz ac voltage range calibration procedure. The second stage is made up of two amplifiers (U305 and U312) each configured for a fixed gain of x10. Overall 2nd stage gains of x1, x10, and x100 are produced by routing the 1st stage output either around, or through one or both amplifiers as shown in the table below.
2nd Stage Gain x1 x10 x100 U306A ON OFF OFF U306B OFF ON ON U306C OFF OFF ON U306D OFF ON OFF U304C OFF OFF ON
The output of the 2nd stage is connected to the rms-to-dc converter stage. Any residual dc offset from the amplifier stages is blocked by capacitor C316. Buffer U307 drives the input to the rms-to-dc converter as well as the frequency comparator (U310A) input. The rms-to-dc converter has two selectable averaging filters (C318 and C318 plus C321) for the analog computer circuit of U308. The two analog averaging filters together with digital filters running in the main CPU implement the three selectable ac filters: slow, medium, and fast. The faster analog filter (using C318) is used for all AC V, AC I, and frequency or period autoranging. The slower analog filter is used only with the slow and medium ac filter choices. In frequency or period measurements, U310A generates a logic signal (FREQIN) for every input zero crossing. The ac sections FREQRNG dc output is measured directly by the main CPUs 10-bit ADC during frequency or period measurements. This lower resolution measurement is sufficient to perform voltage ranging decisions for these functions. The frequency comparator output is disabled during ac voltage and current measurements by U310B forcing U310As input to 15 volts.
135
A-to-D Converter
Unless otherwise noted, components in this discussion are located on the A4 circuit assembly (34970-66504). The schematics are included in Chapter 8 starting on page 234. The analog-to-digital converter (ADC) is used to change dc voltages into digital information (schematic shown on page 9-12). The circuitry consists of an integrator amplifier (U402 and U420), current steering switch U411, resistor network U102E, voltage reference U403, ADC controller U209, and residue ADC U205. The ADC method used is called multislope III. It is based on patented Agilent ADC technology. Multislope III is a charge balancing continuously integrating analog-to-digital converter. The ADC charge balancing algorithm is always running, even when the multimeter is not triggered. The input voltage continuously forces charge onto the integrator capacitors C400 and C401 through U102ER16. Switches U411A and U411B steer fixed positive or negative reference currents onto the integrator capacitor to cancel, or balance, the accumulated input charge. The level shifted (R403 and R406) output of the integrator is checked every 2.66 s by the A1U209 COMP input. Logic state machines in A1U209 control the U411 current steering to continuously seek an approximate 2.5 V level on the integrator amplifier output, FLASH. If the ADC input voltage ADIN is between 15 V, the integrator output (FLASH) will remain within the 0 V to 5 V range of the A1U205 on-chip ADC. An input greater than +15 V may cause the integrator output (U4026) to saturate at about 18 V. An input less than 15 V may cause U402 to saturate with an output of about +18 V. The A1U205 ADC input (FLASH) is clamped to 0 V or 5 V by R405 and CR403 to protect A1U205. The integrator amplifier is formed by U402 and U420. Resistors R420 and R421 affect the amplifier stability. Amplifier oscillation may occur if their values are incorrect. Amplifier U420 improves the offset voltage characteristics of integrator amplifier U402.
136
Each analog-to-digital conversion begins when the multimeter is triggered. The ADC starts by clearing the integrator slope count in A1U209. At the end of the integration period, the slope count is latched. The slope count provides the most significant bits of the input voltage conversion. The least significant bits are converted by the on-chip ADC of CPU A1U205. The instrument precision voltage reference is U403. Resistor R409 provides a stable bias current for the reference zener diode. R408 and CR404 provide a bias to assure that the reference zener biases to +7 V during power up. IC U400A amplifies the voltage reference to +10 V while amplifier U401A inverts the +10 V reference to 10 V. The reference voltages force precision slope currents for the integrating ADC through U102ER17, R18.
137
Switch Modules
In general, all the switch modules share a common module control circuitry. This circuitry is described below. Each module is described in further detail on the following pages.
Switch Module Name 34901A 20 Channel MUX with T/C Compensation 34902A 34903A 34904A 34905A 34906A 34908A 16 Channel Reed MUX with T/C Compensation 20 Channel Actuator 4 X 8 Matrix Switch 50 Dual 4:1 VHF MUX 75 Dual 4:1 VHF MUX 40 Channel Single-Ended MUX with T/C Compensation Page 142 144 146 147 148 148 149
138
In addition to the +5 Volt power supply (Vcc) and ground, the module controller uses four lines for control and communication: RESET, from the Earth Referenced Logic A1U305. RESET is common to all three slots. The module controller performs a reset when this line goes high. Reset conditions vary for each plug-in. SRQ, to the Earth Referenced Logic A1U305. The SRQ line is a wired-OR line that can be driven by any plug-in. Consequently, any module that asserts SRQ (line low), asserts this line in all other slots and at the Earth Reference Logic. DATA IN, from the Floating Logic A1U205 via the opto isolator A1U312. This line is connected in common to all three slots. DATA OUT, from the module controller to the Floating Logic A1U205 via the opto isolator A1U213. This line is a wired-OR line that can be driven by any module. The DATA IN and DATA OUT lines are optically isolated from the floating logic controller. These lines communicate with the Floating Logic using an asynchronous serial bit stream. The serial communications use an 11 bit protocol; a start bit, 8 data bits, an attention bit, and a stop bit. The attention bit is 1 if the 8 data bits are an address/command, or 0 if the 8 data bits modify or provide data for the previously sent command. The module controller uses the hardwired slot-ID bits to decode the serial bit stream address. When the address/command message address matches the slot-ID, the plug-in is selected and responds to the following commands. All other plug-ins will ignore the commands until a new address/command message is received. A 12 MHz crystal, Y101 is the clock for the module controller. The module ID is a four bit pattern set through R106, R107, R108, and R109. The Ferro-electric RAM U150 provides data storage of the relay usage on the module (see page 55) and the thermocouple reference junction temperature corrections (see page 112). Data in and out of U150 is serial. The 34901A, 34902A, and 34908A each have two onboard, solid state temperature sensors, U151 and U152, physically located near the isothermal block at the input connections. The temperature sensors are used as the thermocouple temperature reference.
Module Reference
139
U101 controls the relays on the module using an 8-bit data bus and three control lines. The data lines are latched and applied to the relay drivers. U101 enters a low-power idle mode when inactive. U101 responds when a command is received or when a scheduled reference junction temperature measurement is taken. The relays use a buffered +5 Volt power supply. U101 supplies two drive enable lines (DR_EN and +5NL_EN) that connect Vcc from the digital bus with the relay drive lines through Q101 (+5R or +5NL). To minimize the current through DGND caused by static discharge, the ground return (ZGND) is isolated from the backplane ground through a bead L102.
Relay Drivers
Two types of single-coil relays are used on the switch modules: latching and non-latching. Typical driver configurations are shown below.
140
The non-latching relay contacts are in the set position (closed) when current flows through the coil. When the current is removed, the relay resets (opens). The positive side of the relay coil is connected to +5NL. The negative side of the relay coil is connected to ZGND through a single NPN transistor. The transistor and +5NL must be on for the relay to stay in the set position. The polarity of the current flow through the latching relay coil determines the set (closed) or reset (open) position of the relay contacts. Latching relays are driven by row and column latches or complimentary transistor pairs. To set a relay (close a channel) in the row column driver circuitry, the appropriate row driver PNP transistor is turned on connecting the +5R supply to the positive side of several relay coils and ZGND is applied to the negative side of the desired relay coil through a column driver NPN transistor. To reset a relay, ZGND is applied through an NPN row driver transistor and +5R through a PNP column driver. Tree latching relays are driven by complimentary transistor pairs that steer the current through the relay coil. The +5R supply is only enabled while the relay changes state. The table below shows the times required for the relays to change state.
Switch Module 34901A 34902A 34903A 34904A 34905A/06A 34908A Open 6 ms 0.40 ms 6 ms 6 ms 10 ms 6 ms Close 6 ms 1.25 ms 6 ms 6 ms 15 ms 6 ms
Module Reference
141
34901A
Components in this discussion are located on the A1 circuit assembly (34901-66501). The schematics are included in Chapter 8 starting on page 239. The control circuitry has four groupings of latches, relay drivers and relays. The three control lines (SEL_A, SEL_B, and SEL_C) from the module controller are divided into five control lines by the binary to octal converter U109.
The row latch, U102, and column latches, U103 and U104, control the relays. The row drivers are divided into four groups of set and reset drivers. Each group of row drivers controls five relays. The column drivers operate as a pair. There are ten column drivers each controlling two relays. The analog bus backplane relays are non-latching.
142
The 20 channels are divided into two banks. The banks are combined by closing K422 for voltage and 2-wire Ohms switching. When K422 is open, the banks are electrically independent of each other and this configuration is used for 4-wire Ohms multiplexing where the Ohms current sources are connected to channels 1 through 10 and the sense is obtained from channels 11 through 20. K421 and K423 control the connection to the analog bus for measurements using the Internal DMM.
Module Reference
K422
The current measurement channels are selected by relay K522. Relays K523 and K524 short the inputs when a channel is not selected. Relay K521 makes the connection to the analog bus for measurement by the Internal DMM.
143
34902A
Components in this discussion are located on the A1 circuit assembly (34902-66501). The schematics are included in Chapter 8 starting on page 245. The control circuitry has of two groupings of latches, relay drivers and relays. The 16 voltage and resistance measurement channels are directly driven. Tree switching controls the 2-wire/4-wire Ohms operation and connections to the analog bus.
The reed relays are non-latching and the relay driver and +5R is applied while a channel is closed.
144
The 16 channels are divided into two banks. The banks are combined by closing K327 for voltage and 2-wire Ohms switching. When K327 is open, the banks are electrically independent of each other and this configuration is used for 4-wire Ohms multiplexing where the Ohms current source is connected to channels 1 through 8 and the sense is obtained from channels 9 through 16. Relays K326 and K328 control the connection to the analog bus for measurements using the internal DMM.
Module Reference
145
34903A
Components in this discussion are located on the A1 circuit assembly (34903-66501). The schematics are included in Chapter 8 starting on page 250. The control circuitry has two grouping of latches, relay drivers and relays. The 20 channels are, for control purposes, arranged into 8 rows by 10 columns.
The row latch, U102, and column latches, U103 and U104, control the relays. The row drivers are divided into four groups of set and reset drivers. Each group of row drivers controls five relays. The column drivers operate as a pair. There are ten column drivers each controlling two relays. The 34903A provides 20 channels of Form C switching.
146
34904A
Components in this discussion are located on the A1 circuit assembly (34904-66501). The schematics are included in Chapter 8 starting on page 254. The control circuitry has four groupings of latches, relay drivers and relays divided into 4 rows by 8 columns.
The row latch, U102, and column latch U103, control the relays. The row drivers are divided into four groups of set and reset drivers. Each group of row drivers controls eight relays. The column drivers operate as a pair. There are eight column drivers each controlling four relays. The relays are arranged in 4 rows by 8 columns.
Module Reference
147
34905A/34906A
Components in this discussion are located on the A1 circuit assembly (34905-66501 or 34906-66501). The schematics are included in Chapter 8 starting on page 258. The control circuitry has of two grouping of buffers, relay drivers and relays, one for each multiplexer bank.
Bank1 latch, U102, and Bank2 latch, U103, control the relays. The bank drivers are divided into six groups of set and reset drivers. Each set and reset driver controls one relay. The column drivers operate as a pair. There are six column drivers each controlling a relay. The relays are arranged into two independent banks:
148
34908A
Components in this discussion are located on the A1 circuit assembly (34908-66501). The schematics are included in Chapter 8 starting on page 267. The control circuitry has three grouping of latches, relay drivers and relays. The 40 voltage and resistance measurement channels are, for control purposes, arranged into 8 rows by 10 columns. Tree switching controls bank selection and connections to the analog bus.
Module Reference
The row latch, U102, and column latches, U103 and U104, control the relays. The row drivers are divided into four groups of set and reset drivers. Each group of row drivers controls five relays. The column drivers operate as a pair. There are ten column drivers each controlling two relays. The analog bus backplane relays are non-latching.
149
A single relay is used to switch two input channels. The choice of which channel is connected to the common is performed by relay K422. Channels are paired 20 channels apart (Ch 1 with Ch 21, Ch 2 with Ch 22, etc.) Relay K421 connects the common to the backplane analog bus for use with the internal DMM.
150
Multifunction Module
The 34907A Multifunction module contains two 8-bit digital input/ output ports, a totalizer input, and two 16-bit analog outputs.
Multifunction Control
Components in this discussion are located on the A1 circuit assembly (34907-66501). The schematics are included in Chapter 8 starting on page 261. A simplified block of the module control circuit is shown below.
Module Reference
151
In addition to the +5 Volt power supply (Vcc) and ground, the module controller uses four lines for control and communication: RESET, from the Earth Referenced Logic A1U305. RESET is common to all three slots. The module controller performs a reset when this line goes high. Reset conditions vary for each plug-in. SRQ, to the Earth Referenced Logic A1U305. The SRQ line is a wired-OR line that can be driven by any plug-in. Consequently, any module that asserts SRQ (line low), asserts this line in all other slots and at the Earth Reference Logic. DATA IN, from the Floating Logic A1U205 via the opto isolator A1U312. This line is connected in common to all three slots. DATA OUT, from the module controller to the Floating Logic A1U205 via the opto isolator A1U213. This line is a wired-OR line that can be driven by any module. The DATA IN and DATA OUT lines are optically isolated from the floating logic controller. These lines communicate with the Floating Logic using an asynchronous serial bit stream. The serial communications use an 11 bit protocol; a start bit, 8 data bits, an attention bit, and a stop bit. The attention bit is 1 if the 8 data bits are an address/command, or 0 if the 8 data bits modify or provide data for the previously sent command. The module controller uses the hardwired slot-ID bits to decode the serial bit stream address. When the address/command message address matches the slot-ID, the plug-in is selected and responds to the following commands. All other plug-ins will ignore the commands until a new address/command message is received. A 12 MHz crystal, Y101, is the clock for the module controller. The module ID is a four bit pattern set through RP102. The Ferroelectric RAM U102 provides data storage of the calibration constants for the analog output channels. Data in and out of U102 is serial. 8 data lines, connected to U101 port 1, are used by the digital input and output ports. One of the data lines is used to send serial data to the analog output channels. The totalizer edge count is controlled by U101 P0.6 and read at U101 P3.4. U101 enters a low-power idle mode when inactive. U101 responds when a command is received or when a scheduled alarm scan is needed.
152
Totalizer
Components in this discussion are located on the A1 circuit assembly (34907-66501). The schematics are included in Chapter 8 starting on page 261. A simplified block diagram of the totalizer input is shown below.
5
The totalizer counts signals connected to the COUNT+ and COUNT inputs. Two op-amps, U108A and U108B, are used for input signal conditioning. Comparator U109 determines the signal trigger levels based upon the setting of the jumper at P102. With the P102 jumper in the TTL position, the totalizer counts pulses with TTL trigger levels. With the jumper at P102 in the AC position the trigger level is at zero. The GATE and GATE* input signals control when counting occurs. If no signal is connected, the totalizer counts any changing signal on the inputs. A TTL low on the GATE input or a TTL high on GATE* input will halt counting. Count edge selection is controlled from a U101 port bit (P0.6) through the exclusive OR gate U111. When the P0.6 signal is low, the count increments on the rising edge of the input signal. When the P0.6 signal is high, the count increments on the falling edge of the input signal.
Module Reference
153
Analog Output
Components in this discussion are located on the A1 circuit assembly (34907-66501). The schematics are included in Chapter 8 starting on page 261. A simplified block diagram of the analog output channels is shown below.
Communication with each DAC (U503 and U504) is via three lines: SERSTB, DACCLK, and SERDAT. Each DAC has a voltage output of 3 V. U505 and U506 amplify this voltage to the 12V output. A DC/DC converter is used to provide the 15V supplies to U505 and U506. The 15V supplies also are used at the input of the totalizer. U502 provides the 5 V supply used by the DACs. A line from U101 P0.4 is used to control the output of U510. After a reset or power-up, U510 is held in the shutdown state. U101 turns on the DC/DC converter in response to commands from the main controller. The main controller paces the turn on of the DC/DC converters to ensure that if multiple modules are installed, the backplane power supply is not pulled down by the in-rush current of the DC/DC converters.
154
Digital I/O
Components in this discussion are located on the A1 circuit assembly (34907-66501). The schematics are included in Chapter 8 starting on page 261. A simplified diagram of a digital I/O channel is shown below.
Module Reference
Two stages of latches on the outputs and one set of latches on the inputs provide synchronous 16 bit writes and reads of the digital ports. For digital output, data is written to the upper and lower bytes (U201 and U202) separately, then latched into the output latches U203 and U204 simultaneously. On a digital input data is latched into the input latches U105 and U106 simultaneously. MOSFETs are used to provide the low level output, and 74HC240s are used to provide the high level output. During an output low, a logic high level is applied to the gate of the MOSFET causing it to conduct and creating a low resistance path from the data line to ZGND. In this state the MOSFET is capable of sinking an externally supplied current of up to 400 mA. The blocking diodes, CR301-CR308 and CR401-CR408 prevent any current from sinking into the 74HC240s. During an output high, a logic low level is applied to the gate of the MOSFET turning it off and presenting a high resistance between the data line and ground.
155
The 74HC240s, U205 and U206, provide the output high drive current necessary to maintain a TTL high output level ( 2.4 Vdc) under load. At instrument turn-on, following a reset, and whenever the data lines are being read, the MOSFETs are in the passive high state, and the high output drivers are disabled. The resistor connected between the MOSFETs gate and ZGND holds the gate near ground potential when the module is initially turned-on to ensure that the MOSFET is in the passive high state. The comparators U301, U302, U401 and U402 maintain correct TTL high and low levels by shifting the voltages from the input to compensate for the forward voltage drop of the blocking diode. A reference voltage of +2.1 Vdc (TTL_REF) is applied to the inverting input of the comparator. When the input voltage is in the range of 0 Vdc to +4.3 Vdc the blocking diode is forward biased and its forward voltage drop is added to the applied voltage. For example, when 0 Vdc is applied to the data line, +0.7 Vdc is present on the non-inverting input of the comparator and the comparator output is low. When the input signal level is above 1.4 Vdc, a voltage greater then +2.1 Vdc is applied to the non-inverting input of the comparator causing its output to go high. When the input signal is less than 1.4 Vdc, a voltage less than 2.1 Vdc is applied to the comparators non-inverting input causing its output to go low. This ensures an input voltage < 1.4 Vdc is interpreted as a TTL low level and an input > 1.4 Vdc is interpreted as a TTL high level. The pull-up resistor (connected to the comparators non-inverting input) allows external ground connections and open circuits to be detected. When the data line is grounded, the blocking diode is forward biased applying a +0.7 Vdc level to the comparator, a TTL low. When the data line is allowed to float, the non-inverting input of the comparator pulls up to +5 Vdc, a TTL high. The blocking diode on the output is used for circuit protection. The diode reverse biases when the applied voltage exceeds +4.3 Vdc preventing externally supplied current from being injected into the modules +5V supply line. The MOSFETs have a built in zener diode that conducts at any voltage of approximately 75 Vdc or greater. The zener diode provides protection from external over voltage situations including static electricity.
156
Service
Service
This chapter discusses the procedures involved for returning a failed instrument to Agilent for service or repair. Subjects covered include the following: Operating Checklist, on page 159 Types of Service Available, on page 160 Repackaging for Shipment, on page 161 Cleaning, on page 161 Electrostatic Discharge (ESD) Precautions, on page 162 Surface Mount Repair, on page 162 To Replace the Power-Line Fuse, on page 163 (also depicted on page 42) Troubleshooting Hints, on page 163 Self-Test Procedures, on page 167 Battery Check and Replacement, on page 172 Disassembly, on page 174
158
Operating Checklist
Before returning your instrument to Agilent Technologies for service or repair check the following items:
159
160
Cleaning
Clean the outside of the instrument and plug-in modules with a soft, lint-free, slightly dampened cloth. Do not use detergent. Disassembly is not required or recommended for cleaning.
161
WARNING
162
Troubleshooting Hints
This section provides a brief check list of common failures. Before troubleshooting or repairing the instrument, make sure the failure is in the instrument rather than any external connections. Also make sure that the instrument is accurately calibrated within the last year (see page 63). The instruments circuits allow troubleshooting and repair with basic equipment such as a 612-digit multimeter.
Unit is Inoperative
Verify that the ac power cord is connected to the instrument. Verify that the front-panel On/Standby switch has been pushed. Verify that the power-line fuse is installed and not open (see page 42). The instrument is shipped from the factory with a 500 mAT, 250 V fuse. This is the correct fuse for all line voltages. Verify the power-line voltage setting. See If the Instrument Does Not Turn On on page 42. Remove all plug-in modules to verify that a plug-in module is not causing the failure.
163
Isolating to an Assembly
Remove all plug-in modules to isolate between the instrument and the plug-in modules. Listen for a beep when you press the On/Standby switch. The main controller can operate the beeper even with a display failure. Listen for the fan when you press the On/Standby switch. Fan operation indicates some operation of the main controller and power supplies. Try to operate the instrument from a remote interface. If remote operations are normal, the display should be replaced or repaired. Isolate the internal DMM by removing it. The instrument should operate and pass self-test without the internal DMM installed. Disassembly procedures start on page 174.
164
Power Supplies
Verify the power supplies generated on the 34970-66501 circuit board. The front panel filament voltage, + 5 V backplane and +5 V fan are switched by the On/Standby switch. All other power supplies operate whenever the AC power cord is connected. Warning Exposed Mains Do Not Touch To check the power supplies, remove the instrument cover as shown on page 174. The power supplies can be checked from the bottom of the instrument as shown below. Be sure to use the correct ground point when checking the supplies.
6 Vrms filament
18 V float
3 V battery +
165
Check that the input to the supply voltage regulator is at least 1 V greater than its output. Circuit failures can cause heavy supply loads which may pull down the regulator output voltage. Always check that the power supplies are free of ac oscillations using an oscilloscope. All plug-in modules use the +5 V backplane supply. Some plug-in modules generate their own local power supplies.
166
Self-Test Procedures
Power-On Self-Test
Each time the instrument is powered on, a small set of self-tests are performed. These tests check that the minimum set of logic and measurement hardware are functioning properly. Any plug-in modules installed are verified for two-way communication with the main controller.
Complete Self-Test
Hold down any front panel key for 5 seconds while turning on the power to perform a complete self-test. The instrument beeps when the test starts. If all self-tests pass the display shows PASS for five seconds and the instrument returns to the last measurement function.
Note: The following pages contain a subset of the 34970A error messages. Refer to Chapter 6 in the 34970A Users Guide (p/n 34970-90003) for the complete error message listing.
167
Self-Tests
A complete self-test performs the following tests. A failing test is indicated by the test number and description in the display. 601 Front panel not responding The main CPU A1U205 attempts to establish serial communications with the front panel processor A2U1. During this test, A2U1 turns on all display segments. Communication must function in both directions for this test to pass. If this error is detected during power-up self-test, the instrument will beep. This error is only readable from the remote interface. RAM read/write failed This test writes and reads a 55 and AA checkerboard pattern to each address of RAM. Any incorrect readback will cause a test failure. This error is only readable from the remote interface. A / D sync stuck The main CPU issues an A/ D sync pulse to A1U209 and A1U205 to latch the value in the ADC slope counters. A failure is detected when a sync interrupt is not recognized and a subsequent time-out occurs. A / D slope convergence failed The input amplifier is configured to the measure zero (MZ) state in the 10 V range. This test checks whether the ADC integrator produces nominally the same number of positive and negative slope decisions ( 10%) during a 20 ms interval. Cannot calibrate rundown gain This test checks the nominal gain between the integrating ADC and the A1U205 on-chip ADC. This error is reported if the procedure can not run to completion due to a hardware failure. Rundown gain out of range This test checks the nominal gain between the integrating ADC and the A1U205 on-chip ADC. The nominal gain is check to 10% tolerance. Rundown too noisy This test checks the gain repeatability between the integrating ADC and the A1U205 on-chip ADC. The gain test (606) is performed eight times. Gain noise must be less than 64 LSBs of the A1U205 on-chip ADC.
602
603
604
605
606
607
168
608
Serial configuration readback failed This test re-sends the last 9 byte serial configuration data to all the serial path. The data is then clocked back into A1U209 and compared against the original 9 bytes sent. A failure occurs if the data do not match. DC gain x1 failed This test configures for the 10 V range. The dc amplifier gain is set to X1. The measure customer (MC) input is connected to the internal TSENSE source which produces 0.6 volts. A 20 ms ADC measurement is performed and checked against a limit of 0.6 V 0.3 V. DC gain x10 failed This test configures for the 1 V range. The dc amplifier gain is set to X10. The measure customer (MC) input is connected to the internal TSENSE source which produces 0.6 volts. A 20 ms ADC measurement is performed and checked against a limit of 0.6 V 0.3 V. DC gain x100 failed This test configures for the 100 mV range. The dc amplifier gain is set to X100. The measure customer (MC) input is connected to the internal TSENSE source which produces 0.6 volts. A 20 ms ADC measurement is performed and checked for a + overload response. Ohms 500 nA source failed This test configures to the 10 V dc range with the internal 10 M 100:1 divider A4U102 connected across the input. the 500 nA Ohms current source is connected to produce a nominal 5 V signal. A 20 ms ADC measurement is performed and the result is checked against a limit of 5 V 1 V. Ohms 5 A source failed This test configures the 10 V range with the internal 10 M 100:1 divider A4U102 connected across the input. The 5 A current source is connected. The compliance limit of the current source is measured. A 20 ms ADC measurement is performed and the result is checked against a limit of 7.5 V 3 V. DC 300V zero failed This test configures the 300 V dc range with no input applied. A 20 ms ADC measurement is performed and the result is checked against a limit of 0V 5 mV.
609
610
611
612
613
614
169
615
Ohms 10 A source failed This test configures the 10 V range with the internal internal 10 M 100:1 divider A4U102 connected across the input. The 10 A current source is connected. A 20 ms ADC measurement is performed and the result is checked against a limit of 7.5 V 3 V. DC current sense failed This test configures the 1 A dc rage and function. A 20 ms ADC measurement is performed and the result is checked against a limit of 0 A 5 A. This test confirms that the dc current sense path is functional. Ohms 100 A source failed This test configures the 10 V range with the internal 10 M 100:1 divider A4U102 connected across the input. The 100 A current source is connected. The compliance limit of the current source is measured. A 20 ms ADC measurement is performed and the result is checked against a limit of 5 V 1 V. DC high voltage attenuator This test configures to the 300 Vdc range. the 500 nA ohms current source is connected to produce a nominal 5 V signal. A 20 ms ADC measurement is performed and the result is checked against a limit of 10 mV to 70 mV at the output of the rms-to-dc converter. Ohms 1 mA source failed This test configures the 10 V range with the internal 1 0 M 100:1 divider A4U102 connected across the input. The 1 mA current source is connected. A 20 ms ADC measurement is performed and the result is checked against a limit of 7 V 3.5 V. AC rms zero failed This test configures to the 100 mV ac range with the ac input grounded through A4K103. The internal residual noise of the ac section is measured and checked against a limit of 10 mV to 70 mV at the output of the rms-to-dc converter. AC rms full scale failed This test configures for the 100 mV ac range. The 1 mA ohms current source is switched on the charge the ac input capacitor A4C301. This produces a pulse on the output of the rms-to-dc converter which is sampled 100 ms after the current is applied. A 20 ms A/D measurement is performed and checked against a limit of 10 V 8.5 V into the ADC.
616
617
618
619
620
621
170
622
Frequency counter failed This test configures for the 100 mV ac range. This test immediately follows test 621. With A4C301 holding charge from test 621 the ac input is now switched to ground through A4K103. This produces a positive pulse on the input to the frequency comparator A4U310. While C301 discharges, the ENAB FREQ bit is toggled four times to produce a frequency input to the counter logic in A1U205. A failure occurs if the counter can not measure the frequency input. Cannot calibrate precharge This test configures to the 100 V dc range with no input. The ADC is configured for 200 ms measurements. The A1U205 pulse width modulated (PWM) DAC output (C224) is set to about 4 volts. A reading is taken in with A4U101 in the MC state. A second reading is taken in the PRE state. The precharge amplifier voltage offset is calculated. The A1U205 DAC output is set to about 1.5 volts and the precharge offset is measured again. The gain of the offset adjustment is calculated. This test assures a precharge amplifier offset is achievable. Unable to sense line frequency This test checks that the LSENSE logic input to A1U205 is toggling. If no logic input is detected, the meter will assume a 50 Hz line operation for all future measurements. I/O processor did not respond This test checks that communications can be established between A1U205 and A1U305 through the optically isolated (A1U213 and A1U214) serial data link. Failure to establish communication in either direction will generate an error. If this condition is detected at power-on self-test, the instrument will beep and the error annunciator will be on. I/O processor failed self-test A failure occurred when the earth referenced processor, AU305, executed an internal RAM test.
623
624
625
626
171
3 V battery
172
Battery
Battery +
6
To Replace the Battery
1 2 3 Remove AC line power. Remove the cover (see page 174). If installed, remove the Internal DMM (A4) assembly (see page 176). Turn the instrument over and unsolder the three battery terminals from the bottom of the circuit board. Use proper through-hole soldering techniques and equipment. Remove the battery and dispose of properly. Install, solder, and verify the new battery.
173
Disassembly
The following tools are recommended for disassembly. T15 Torx driver (all screws) 11 mm nut driver (front-panel disassembly) 5 mm nut driver (rear-panel connectors) Tighten the fan screws to a maximum of 6 in/lbs (0.68 newton/meter). WARNING SHOCK HAZARD. Only service-trained personnel who are aware of the hazards involved should remove the instrument covers. Dangerous voltages may be encountered with the instrument covers removed. To prevent damage to the fan, do not over tighten the fan screws.
CAUTION
174
General Disassembly
175
176
Front-Panel Disassembly
Note: When reassembling the front panel, be sure to route the front-panel cable as shown above. Do not allow the front-panel cable to touch the digital ribbon cable. 177
178
Needle-nose Pliers
179
180
Replaceable Parts
Replaceable Parts
This chapter contains information ordering replacement parts for your instrument. The parts lists are divided into the following groups. 34970A Mainframe, on page 183 34970-66501 Main PC Assembly (A1), on page 184 34970-66502 Front-Panel and Keyboard PC Assembly (A2), on page 189 34970-66503 Backplane PC Assembly (A3), on page 190 34970-66504 Internal DMM PC Assembly (A4), on page 191 34901A 20-Channel Multiplexer, on page 196 34902A 16-Channel Multiplexer, on page 200 34903A 20-Channel Actuator, on page 202 34904A 4x8 Matrix, on page 204 34905A/34906A RF Multiplexer, on page 207 34907A Multifunction Module, on page 209 34908A 40-Channel Multiplexer, on page 213
Manufacturers List, on page 216 Parts are listed in alphanumeric order according to their schematic reference designators. The parts lists include a brief description of the part with applicable Agilent part numbers and manufacturer part number.
182
34970A Mainframe
Reference Designator A1 A2 A3 A4
CBL1 CBL2 CBL3 CVR1 FRM1 HDW1 HDW2 MNL1 MP1 MP2 MP3 MP4 MP5 MP6 MP7 MP8 MP9 MP10 MP11 MP12 SCW1-SCW7 SCW8-SCW9 SCW10 SHD1 SHD2 SHD3 T1
Part Number 34970-66501 34970-66502 34970-66503 34970-66504 34970-80010 34970-61606 RS232-61601 8120-1378 34970-84101
34970-80101 0535-0154 3050-1941 34970-90101 34401-45011 34401-86010 34970-00101 34970-00102 34970-40201 34970-44111 34970-49321 34970-68501 34970-86201 34970-87401 34970-88001 34970-88301 0515-0433 0624-0751 0624-0078 34970-00601 34970-00602 34970-00603 9100-5608
Qty
1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 2 1 1 1 1 1 1 7 2 1 1 1 1 1
Part Description
PCA-DIGITAL BD PCA-DISPLAY BD PCA-BACKPLANE BD PCA-DMM BD Internal DMM Field Kit CABLE, K-TYPE TC, SCRWDRVR CABLE, RS232, 9 PIN CBL-U.S. COVER-SHEET METAL CHASSIS NUT HEX 11MM X2T WASHER-FL NM 7/16 IN .353-IN-ID MANUAL SET HANDLE-FRONT KIT-BUMPERS/COVER CARD CAGE, LEFT CARD CAGE, RIGHT FRNT PANEL ASSY COVER PLATE, PLASTIC WINDOW/FRONT FAN PWR MOD W/FUSE KNOB KEYPAD BEZEL-REAR, MLD SCRPHM4.0X08TXSC SCR 6-19X1/2TORX GROUND SCREW 6-32 .375-IN-LG POZI SHIELD-XFMR SHIELD-AC LINE SHIELD-DMM XFMR-PWR 100/120/220/240V
Mfr Code 02362 02362 02362 02362 02362 02362 02362 04940 02361
02361 11239 05227 02362 02362 02362 02362 02362 02362 02362 02362 02362 02362 02362 02362 02362 02361 05610 05525 02361 02361 02362 02859
183
Qty
1 3 1 39 3 2 2
Part Description
BATTERY 3V 850A-HR LI MANGANESE DIOXIDE CAP-FXD 2200uF +-20% 25 V AL-ELCTLT CAP-FXD 33uF +-20% 20 V TA CAP 0.1UF 50V 10% X7R 0805 CAP-FXD 2200uF +-20% 25 V AL-ELCTLT CAP-FXD 22uF +-20% 20 V TA CAP-FXD 1000uF +-20% 50 V AL-ELCTLT CAP-FXD 1uF +-20% 35 V TA CAP-FXD 1000uF +-20% 50 V AL-ELCTLT CAP-FXD 1uF +-20% 35 V TA CAP-FXD 0.01uF +-10% 50 V CER X7R CAP-FXD 2200uF +-20% 25 V AL-ELCTLT CAP 0.1UF 50V 10% X7R 0805 CAP-FXD 22uF +-20% 20 V TA CAP 0.1UF 50V 10% X7R 0805 CAP 0.1UF 50V 10% X7R 0805 CAP-FXD 1uF +-10% 10 V CER CAP-FXD 1000pF 50 V CAP 0.1UF 50V 10% X7R 0805 CAP 0.1UF 50V 10% X7R 0805 CAP 0.1UF 50V 10% X7R 0805 CAP 0.1UF 50V 10% X7R 0805 CAP-FXD 0.01uF +-10% 50 V CER X7R CAP 0.1UF 50V 10% X7R 0805 CAP-FXD 4.7uF +-20% 10 V TA CAP-FXD 0.01uF +-10% 50 V CER X7R CAP-FXD 47pF +-5% 50 V CER C0G CAP-FXD 47uF 10 V TA CAP-FXD 1000pF 50 V CAP 0.1UF 50V 10% X7R 0805 CAP 0.1UF 50V 10% X7R 0805 CAP-FXD 47pF +-5% 50 V CER C0G CAP-FXD 3300pF 50 V FIXED CAPACITOR; 22PF 50 VOLTS CAP-FXD 47pF +-5% 50 V CER C0G CAP-FXD 10uF +-20% 35 V TA CAP-FXD 1000pF 50 V
1 6
1 4 1
1 1 4
184
Qty
Part Description
CAP 0.1UF 50V 10% X7R 0805 CAP-FXD 1000pF 50 V CAP-FXD 0.01uF +-10% 50 V CER X7R CAP 0.1UF 50V 10% X7R 0805 CAP-FXD 0.01uF +-10% 50 V CER X7R
4 2 2 1 12
DIODE, 3A, 400V, 75NS,NON-PHOTOSENSITIVE DIODE-FW BRDG 400V 1A DIODE-ZNR 6.2V 5% PD=1.5W IR=5UA DIODE-ZNR 7.5V 5% TO-236 (SOT-23) DIODE-FW BRDG 400V 1A DIODE-DUAL 70V 100MA T0-236AA DIODE-DUAL 70V 100MA T0-236AA DIODE-DUAL 70V 100MA T0-236AA DIODE-DUAL 70V 100MA T0-236AA CORE-SHLD BEAD HEAT SINK SGL TO-220-CS HEAT SINK SGL TO-220-CS CORE-SHLD BEAD CORE-SHLD BEAD CORE-MAGNETIC CORE-SHLD BEAD CORE-SHLD BEAD CORE-SHIELDING BEAD CORE-SHLD BEAD CONN-POST TYPE 2.0-PIN-SPCG 12-CONT CONN FRCC VERT MALE 26PIN CONN-RECT MICRORBN 24-CKT 24-CONT WASHER- NO. 10 .194-IN-ID .294-IN-OD HEAT SINK WASHER- NO. 10 .194-IN-ID .294-IN-OD STANDOFF-HEX .255-IN-LG 6-32-THD STANDOFF-HEX .255-IN-LG 6-32-THD CONN-POST TYPE .156-PIN-SPCG 3-CONT CONN-POST TYPE .156-PIN-SPCG 8-CONT CONN-POST TYPE 2.5-PIN-SPCG-MTG-END
15 3
1 1 1 2 1 2
1 1 1
185
Qty
1 1 1 1 1 1 1 6 2
Mfr Code CONNECTOR-SGL CONT PIN 1.14-MM-BSC-SZ SQ 01136 CABLE, DGTL-BKPLN 03418 CABLE, RS232-DGTL 04726 Part Description
TRANSISTOR-MOSFET DUAL P-CHAN E-MODE SI TRANSISTOR NPN SI TO-236AA PD=350MW FET NMOS 2X S0BN 30V 1.5A R MMDF1N05 TRANSISTOR PNP SI TO-261AA (SOT-223) RESISTOR 0 +-5% .1W TKF TC=0+-300 RESISTOR 42.2K +-1% .1W TKF TC=0+-100 RESISTOR 0 +-5% .1W TKF TC=0+-300 RESISTOR 42.2K +-1% .1W TKF TC=0+-100 RESISTOR 10K +-1% .1W TKF TC=0+-100 RESISTOR 237 +-1% .1W TKF TC=0+-100 RESISTOR 3.32K +-1% .1W TKF TC=0+-100 RESISTOR 0 +-5% .1W TKF TC=0+-300 RESISTOR 237 +-1% .1W TKF TC=0+-100 RESISTOR 3.16K +-1% .1W TKF TC=0+-100 RESISTOR 0 +-5% .1W TKF TC=0+-300 RESISTOR 10K +-1% .1W TKF TC=0+-100 RESISTOR 100K +-1% .1W TKF TC=0+-100 RESISTOR 1K +-1% .1W TKF TC=0+-100 RESISTOR 100K +-1% .1W TKF TC=0+-100 RESISTOR 0 +-5% .1W TKF TC=0+-300 RESISTOR 14.7K +-1pct .1W TKF TC=0+-100 RESISTOR 100K +-1% .1W TKF TC=0+-100 RESISTOR 46.4K +-1% .1W TKF TC=0+-100 RESISTOR 1K +-1% .1W TKF TC=0+-100 RESISTOR 100 +-1% .1W TKF TC=0+-100 RESISTOR 10K +-1% .1W TKF TC=0+-100 RESISTOR 215K +-1% .1W TKF TC=0+-100 RESISTOR 10M +-5% .1W TKF TC=0+-100 RESISTOR 100 +-1% .1W TKF TC=0+-100 RESISTOR 100K +-1% .1W TKF TC=0+-100 RESISTOR 5.62K +-1% .1W TKF TC=0+-100 RESISTOR 100 +-1% .1W TKF TC=0+-100 RESISTOR 1K +-1% .1W TKF TC=0+-100 RESISTOR 1K +-1% .1W TKF TC=0+-100 RESISTOR 10K +-1% .1W TKF TC=0+-100 02037 02237 02037 06121 06337 06337 06337 06337 06337 06337 00746 06337 06337 02995 06337 06337 02995 06337 02995 06337 05524 02995 06337 06337 06337 06337 06337 05524 06337 02995 06337 06337 06337 06337 06337
9 2 1
31 17
1 13 19 2 1
186
Reference Designator R222 R251 R252-R254 R255-R259 R260 R261 R262 R270 R271-R273 R275 R277 R301-R302 R303 R304 R305-R309 R310 R311 R312 R313 R314 R315 R330-R331 R340 R350 R351-R358 R360-R362 R370 R371-R372 R373 R375 R380 R401 R403-R423
U101 U102 U103 U104 U105 U106
Part Number 0699-3034 0699-3058 0699-3034 0699-3058 0699-3053 0699-2965 0699-3053 0699-3001 0699-2977 0699-2983 0699-2983 0699-3034 0699-3053 0699-3034 0699-3058 0699-3034 0699-3050 0699-3034 0699-3077 0699-3035 0699-3034 0699-2977 0699-2965 0699-3058 0699-2965 0699-3058 0699-2977 0699-2965 0699-3051 0699-3051 0699-3034 0699-2643 0699-3053
1826-2794 1826-2264 1826-1597 1826-3044 1826-0393 1826-0527
Qty
Part Description
RESISTOR 1K +-1% .1W TKF TC=0+-100 RESISTOR 100 +-1% .1W TKF TC=0+-100 RESISTOR 1K +-1% .1W TKF TC=0+-100 RESISTOR 100 +-1% .1W TKF TC=0+-100 RESISTOR 100K +-1% .1W TKF TC=0+-100 RESISTOR 46.4K +-1% .1W TKF TC=0+-100 RESISTOR 100K +-1% .1W TKF TC=0+-100 RESISTOR 215K +-1pct .1W TKF TC=0+-100 RESISTOR 681 +-1% .1W TKF TC=0+-100 RESISTOR 5.62K +-1% .1W TKF TC=0+-100 RESISTOR 5.62K +-1% .1W TKF TC=0+-100 RESISTOR 1K +-1% .1W TKF TC=0+-100 RESISTOR 100K +-1% .1W TKF TC=0+-100 RESISTOR 1K +-1% .1W TKF TC=0+-100 RESISTOR 100 +-1% .1W TKF TC=0+-100 RESISTOR 1K +-1% .1W TKF TC=0+-100 RESISTOR 100K +-1% .1W TKF TC=0+-100 RESISTOR 1K +-1% .1W TKF TC=0+-100 RESISTOR 1M +-1% .1W TKF TC=0+-100 RESISTOR 1.47K +-1% .1W TKF TC=0+-100 RESISTOR 1K +-1% .1W TKF TC=0+-100 RESISTOR 681 +-1% .1W TKF TC=0+-100 RESISTOR 46.4K +-1% .1W TKF TC=0+-100 RESISTOR 100 +-1% .1W TKF TC=0+-100 RESISTOR 46.4K +-1% .1W TKF TC=0+-100 RESISTOR 100 +-1% .1W TKF TC=0+-100 RESISTOR 681 +-1% .1W TKF TC=0+-100 RESISTOR 46.4K +-1% .1W TKF TC=0+-100 RESISTOR 10K +-1% .1W TKF TC=0+-100 RESISTOR 10K +-1% .1W TKF TC=0+-100 RESISTOR 1K +-1% .1W TKF TC=0+-100 RESISTOR 0 +-5% .1W TKF TC=0+-300 RESISTOR 100K +-1% .1W TKF TC=0+-100 IC V RGLTR-FXD-POS 4.85/5.15V 3-TO-220 IC PWR MGT-UND-V-SEN 8 PINS P-SOIC PKG IC PWR MGT-V-REG-FXD-POS 4.85/5.15V IC PWR MGT-VS-SUPVR/CONT 16 PINS P-SOIC IC PWR MGT-V-REG-ADJ-POS 1.2/37V 3 PINS IC PWR MGT-V-REG-ADJ-NEG 1.2/37V 3 PINS
1 1 1
Mfr Code 06337 06337 06337 06337 02995 06337 02995 06337 06337 06337 06337 06337 02995 06337 06337 06337 06337 06337 06337 06337 06337 06337 06337 06337 06337 06337 06337 06337 06337 06337 06337 06337 02995
03406 02037 03406 03285 01698 03406
1 1 1 1 1 1
187
Reference Designator U107 U150 U201 U204 U205 U209 U212 U213-U214 U215 U220 U301 U302 U303 U304 U305 U306 U307 U308 U309 U310 U311 U312 U320 U401 U402-U405 U410
Part Number 1826-1572 1826-2817 1818-6821 1813-0827 1821-1479 1821-2271 1820-5937 1990-1552 9164-0173 1821-3433 0410-4009 1821-0055 1990-1552 1820-7312 34970-88803 1820-7312 1820-6863 1820-6823 1822-0639 1820-6175 1820-6176 1990-1552 1820-7312 34970-88807 1818-8796 1818-5917
Qty
1 1 1 1 1 1 1 4 1 1 1 1 3 1 1 1 1 1 1
Part Description
IC COMPARATOR PRCN DUAL 8 PIN PLSTC-SOIC IC PWR MGT-V-REF-FXD 4.975/5.025V 3 PINS FRAM SERIAL 4K FMZ4C04-S CLK-OSC-XTAL STD 12.000-MHZ 0.01% IC-80C196 ASIC 0.8U GATE ARRAY PERIPH/MEM CONTROL IC FF CMOS/AC D-TYPE POS-EDGE-TRIG OPTO-ISOLATOR LED-IC GATE IF=10MA-MAX ALARM-AUDIBLE PIEZO ALARM PIN TYPE; 25V 7S14-SINGLE GATE, INVERTER, SCHMIT TRIGTNY CERO-RES 12MHZ +1-0.8% IC SCHMITT-TRIG CMOS/ACT NAND QUAD 2-INP OPTO-ISOLATOR LED-IC GATE IF=10MA-MAX IC SCHMITT-TRIG CMOS/ACT INV HEX PROG PART LOWER LEVEL 1821-1876 IC SCHMITT-TRIG CMOS/ACT INV HEX IC-UART WITH FIFO IC INTERFACE DRVR/RCVR CMOS TPL -888-BIT IC HPIB CONTROLLER IC-INTERFACE XCVR BIPOLAR BUS OCTL IC-INTERFACE XCVR BIPOLAR BUS OCTL OPTO-ISOLATOR LED-IC GATE IF=10MA-MAX IC SCHMITT-TRIG CMOS/ACT INV HEX OTP - PROG 1818-5589 (Rev. 13) IC 1M-BIT SRAM 70-NS CMOS 32-SOP IC 256K-BIT SRAM 70-NS CMOS
1 4 1
Mfr Code 02910 03406 14543 12768 03811 03677 03406 01542 09939 03406 00830 02037 01542 02037 02362 02037 03406 02037 03677 01698 01698 01542 02037 02362 00039 06916
SOCKET-IC-PLCC 44-CONT SQUARE J-LEAD SOCKET-IC-PLCC 44-CONT SQUARE J-LEAD CRYSTAL-QUARTZ 32.768 KHZ CRYSTAL-QUARTZ 3.6864 MHZ
1 1
188
Qty
1 1 11 1 1 1 1 1 1 1 2 4 3
Mfr Code CAP-FXD 1uF +-20% 35 V TA 00039 CAP-FXD 10uF +-20% 35 V TA 05524 CAP 0.1UF 50V 10% X7R 0805 02010 DIODE-ZNR 6.2V 5% TO-236 (SOT-23) 02037 INDICATOR PANEL, VACUM FLOURESCENT DSPLY 11908 Part Description
SHIELD, ESD TEMP PART FOR ASSY34970-66502 CORE-MAGNETIC CABLE, DISPLAY RESISTOR 26.1K +-1% .1W TKF TC=0+-100 RESISTOR 2.61K +-1% .1W TKF TC=0+-100 RESISTOR 100K +-1% .1W TKF TC=0+-100 RESISTOR 10K +-1% .1W TKF TC=0+-100 RESISTOR 100 +-1% .1W TKF TC=0+-100 RESISTOR 10K +-1% .1W TKF TC=0+-100 PGMD 1820-8905 8 BIT MCU W/4K EPROM IC PWR MGT-V-REG-FXD-POS 4.8/5.2V 3 PINS CERO-RES 12MHZ +1-0.8% IC-INTERFACE DRVR BIPOLAR DISPLAY IC COMPARATOR LP QUAD 14 PIN PLSTC-SOIC IC 34064 ROTARY ENCODER 24POS TH-VERT 02361 06352 02632 06337 06337 02995 06337 06337 06337 02632 02037 00830 01698 02037 02037 03744
1 1 1 2 1 1 1
189
Qty
1 2 1 2 3 1 1 1 1 1 2
Part Description
CAP-FXD 0.01uF +-10% 500 V CER X7R TUBE-ELECTRON SURGE V PTCTR CLIP BAT
INDUCTOR 10NH +-10% 2.8W-MMX3.4LG-MM 02366 CONN DIN VERT MALE 48PIN CONN-POST TYPE .100-PIN-SPCG 6-CONT CONN DIS RA M SPIN SMC BACKPLANE CABLE RESISTOR 1M +-1% .125W TKF TC=0+-100 RESISTOR 220 5% AXIA1 .5W TC=0+200PPM DIO, MOV, OPERV=230VRMS, 17J, CU4032K230GK1 02010 03418 03418 00003 04935 05524 06121
190
Part Description
CAP-FXD 470pF 630 V POLYP-FL CAP-FXD 220pF 630 V POLYP-FL CAP-FXD 0.1uF 25 V CAP-FXD 1000pF 50 V CAP-FXD 100pF +-5% 50 V CER C0G CAP-FXD 0.01uF 50 V CAP-FXD 0.1uF 25 V CAP-FXD 100pF +-5% 50 V CER C0G CAP-FXD 0.1uF 25 V CAP-FXD 0.1uF 25 V CAP-FXD 0.1uF 25 V CAP-FXD 0.01uF 50 V CAP-FXD 1000pF 50 V CAP-FXD 0.01uF 50 V CAP 220PF 5% 50V CAP-FXD 0.1uF 25 V CAP 0.22uF 400 V CAP-FXD 1.8pF +-13.89% 1.5 kV CER C0G CAP-FXD 680pF +-5% 50 V CER C0G CAP-FXD 6.8pF 50 V CAP-FXD 470pF +-5% 50 V CER C0G CAP-FXD 5.6pF +-8.93% 50 V CER C0G CAP-FXD 100pF +-5% 50 V CER C0G CAP-FXD 0.1uF 25 V CAP-FXD 68pF +-5% 50 V CER C0G CAP-FXD 0.1uF 25 V CAPACITOR-FXD .22UF +-10% 63VDC CAP. FIXED. CER./CHIP CAPACITOR-FXD .22UF +-10% 63VDC CAP-FXD 0.1uF 25 V CAPACITOR-FXD 1UF +-10% 50VDC CAP-FXD 22uF +-20% 35 V TA CAP-FXD 33pF +-5% 50 V CER C0G CAP-FXD 1000pF 50 V CAP-FXD 33pF +-5% 50 V CER C0G CAP-FXD 68uF +-20% 10 V TA CAP-FXD 0.01uF 50 V CAP 220PF 5% 50V
4 1 1 1 1 1 1
1 2 1
1 2 2
Mfr Code 02995 02995 02010 02010 03292 02010 02010 03292 02010 02010 02010 02010 02010 02010 02010 02010 05176 02010 02010 02010 02010 02010 03292 02010 12473 02010 05524 02010 05524 02010 10881 12340 02010 02010 02010 12340 02010 02010
191
Reference Designator C403-C404 C407-C408 C410 C441-C442 C443 C460 CR103 CR110-CR113 CR115 CR201 CR202 CR203 CR302-CR303 CR304-CR307 CR401-CR402 CR403-CR404 K102-K104 L101
L102 L106 L110-L111 L401-L402 L404 L405-L407 P100 P101 Q104-Q105 Q150-Q153 Q201 Q202 Q203-Q210 Q211 Q301
Part Number 0160-6497 0160-6497 0160-6497 0160-6736 0699-2963 0160-6731 1902-1565 1901-1607 1902-1565 1902-1565 1901-1378 1902-1592 1906-0291 1902-1541 1902-1541 1906-0291 0490-1896 9140-1244
9140-1238 9140-1244 9140-1238 9170-1431 9170-1506 9170-1584 34970-61604 34970-61603 1855-0752 1854-1014 1855-0752 1855-1671 1853-0727 1855-1672 1855-0800
Qty
Part Description
CAP-FXD 0.1uF 25 V CAP-FXD 0.1uF 25 V CAP-FXD 0.1uF 25 V CAP-FXD 0.01uF 50 V RESISTOR 121 +-1pct .1W TKF TC=0+-100 CAP-FXD 1000pF 50 V DIODE-ZNR 4.7V 5% TO-236 (SOT-23) DIODE-PWR RECT 400V DO-214AB DIODE-ZNR 4.7V 5% TO-236 (SOT-23) DIODE-ZNR 4.7V 5% TO-236 (SOT-23) DIODE; HV RECTIFIER 1.6KV DIODE-ZNR 5.1V 5% TO-236 (SOT-23) DIODE-DUAL 70V 100MA T0-236AA DIODE-ZNR 3.3V 5% TO-236 (SOT-23) DIODE-ZNR 3.3V 5% TO-236 (SOT-23) DIODE-DUAL 70V 100MA T0-236AA RLY-2C 2A 220V 60WATT 3V-COIL LATCHING INDUCTOR 1MH +-5% 3.4W-MMX4.8LG-MM Q=30 INDUCTOR 10UH +-5% 2.8W-MMX3.4LG-MM Q=30 INDUCTOR 1MH +-5% 3.4W-MMX4.8LG-MM Q=30 INDUCTOR 10UH +-5% 2.8W-MMX3.4LG-MM Q=30 CORE-SHLD BEAD CORE-MAGNETIC CORE-SHLD BEAD CABLE, DMM-DGTL CABLE, DMM-BKPLN TRANSISTOR J-FET N-CHAN D-MODE TO-236AA TRANSISTOR NPN SI TO-236AA PD=350MW TRANSISTOR J-FET N-CHAN D-MODE TO-236AA TRANSISTOR J-FET P-CHAN D-MODE TO-92 TRANSISTOR PNP SI SOT-23 (TO-236AB) TRANSISTOR J-FET N-CHAN D-MODE TO-92 TRANSISTOR MOSFET N-CHAN E-MODE TO-252AA
3 1 4 1 1 1 4 5
3 2 3
Mfr Code 02010 02010 02010 02010 00746 02010 02910 04733 02910 02910 04504 02037 02910 02037 02037 02910 12921 05524
02366 05524 02366 06352 06352 09808 04726 02362 03406 04200 03406 02237 02237 02237 02037
2 1 3 1 1 2 4
1 8 1 1
192
Reference Designator R103 R104 R105 R112-R113 R114-R117 R118 R119 R121 R122 R123 R124 R126-R127 R130 R131 R150 R151 R170 R173 R183 R184 R196 R197 R198 R201 R202 R203-R206 R207 R290 R301-R302 R303 R304 R305 R306 R307 R308 R309 R310 R311-R312 R313 R314
Part Number 0699-1380 0699-4821 0699-3406 0699-3053 0699-2973 0699-1380 0699-2973 0699-5049 0699-1329 0699-4845 0699-3046 0699-2986 0699-3067 0699-2986 0699-3051 0699-3029 0699-4821 0699-3406 0699-3406 0699-4821 0699-3051 0699-3049 0699-3034 0699-3404 0699-4416 0699-1332 0699-3038 0699-1374 0699-2469 0699-1307 0699-0481 0699-1374 0699-1423 0699-1374 0699-1423 0699-1329 0699-2973 0699-1412 0699-1380 0699-1398
Qty
3 3 3 2 12
1 2 1 1 2 1 2 1
1 1 1 1 4 1 3 2 1 1 3
3 5
Mfr Code RESISTOR 3.16K +-1% .125W TKF TC=0+-100 04935 RESISTOR 75K 5% 2512 200V 1W TC=200 05524 RESISTOR 24K +-5% 1W TKF TC=0+-200 05524 RESISTOR 100K +-1% .1W TKF TC=0+-100 02995 RESISTOR 215 +-1% .1W TKF TC=0+-100 06337 RESISTOR 3.16K +-1% .125W TKF TC=0+-100 04935 RESISTOR 215 +-1% .1W TKF TC=0+-100 06337 RESISTOR 5 +-1pct .125W TF TC=0+-20 05524 RESISTOR 6.19K +-1% .125W TKF TC=0+-100 04935 RESISTOR .10 1% RES3F 2W TC=25PPM/DEGC 05524 RESISTOR 6.19K +-1% .1W TKF TC=0+-100 06337 RESISTOR 21.5K +-1% .1W TKF TC=0+-100 06337 RESISTOR 14.7K +-1% .1W TKF TC=0+-100 06337 RESISTOR 21.5K +-1% .1W TKF TC=0+-100 06337 RESISTOR 10K +-1% .1W TKF TC=0+-100 06337 RESISTOR 316 +-1% .1W TKF TC=0+-100 06337 RESISTOR 75K 5% 2512 200V 1W TC=200 05524 RESISTOR 24K +-5% 1W TKF TC=0+-200 05524 RESISTOR 24K +-5% 1W TKF TC=0+-200 05524 RESISTOR 75K 5% 2512 200V 1W TC=200 05524 RESISTOR 10K +-1% .1W TKF TC=0+-100 06337 RESISTOR 8.25K +-1% .1W TKF TC=0+-100 06337 RESISTOR 1K +-1% .1W TKF TC=0+-100 06337 RESISTOR 400K +-1% .125W TF TC=0+-2 06337 RESISTOR 40K +-1% .5W MF TC=0+-.0031 05524 RESISTOR 196K +-1% .125W TKF TC=0+-100 04935 RESISTOR 2.37K +-1% .1W TKF TC=0+-100 06337 RESISTOR 1.78K +-1% .125W TKF TC=0+-100 04935 RESISTOR 500K +-0.25% .25W TF TC=0+-5 09454 RESISTOR 1.96K +-0.1% .1W TF TC=0+-5 02995 RESISTOR 200K +-1% .1W TF TC=0+-10 02995 RESISTOR 1.78K +-1% .125W TKF TC=0+-100 04935 RESISTOR 215 +-1% .125W TKF TC=0+-100 04935 RESISTOR 1.78K +-1% .125W TKF TC=0+-100 04935 RESISTOR 215 +-1% .125W TKF TC=0+-100 04935 RESISTOR 6.19K +-1% .125W TKF TC=0+-100 04935 RESISTOR 215 +-1% .1W TKF TC=0+-100 06337 RESISTOR 75K +-1% .125W TKF TC=0+-100 04935 RESISTOR 3.16K +-1% .125W TKF TC=0+-100 04935 RESISTOR 21.5K +-1% .125W TKF TC=0+-100 04935 Part Description
193
Reference Designator R315 R316 R317 R318 R319 R320 R321 R322 R323-R324 R325-R326 R327 R398-R399 R403 R405 R406 R407-R408 R409 R420 R421 R422 R430 R440 R441 R442 R450-R451 R460-R461 RV102
SCR101 SHD101 U101 U102 U103 U104 U105 U106 U110 U150 U153
Part Number 0699-1327 0699-1423 0699-1406 0699-1318 0699-1398 0699-1427 0699-1382 0699-1412 0699-1398 0699-2973 0699-1398 0699-1391 0699-1391 0699-1380 0699-1330 0699-1318 0699-1372 0699-1389 0699-1318 0699-1360 0699-1503 0699-1406 0699-1394 0699-2127 0699-2973 0699-2973 0837-0320
0515-0433 34970-00603 1SK6-0001 1NB4-5035 1826-2420 1855-0864 1826-2558 1826-1925 1826-1810 1820-8937 1826-2420
Qty
1 2 4 1 1
Part Description
RESISTOR 1M +-1% .125W TKF TC=0+-100 RESISTOR 215 +-1% .125W TKF TC=0+-100 RESISTOR 42.2K +-1% .125W TKF TC=0+-100 RESISTOR 1K +-1% .125W TKF TC=0+-100 RESISTOR 21.5K +-1% .125W TKF TC=0+-100 RESISTOR 316 +-1% .125W TKF TC=0+-100 RESISTOR 3.83K +-1% .125W TKF TC=0+-100 RESISTOR 75K +-1% .125W TKF TC=0+-100 RESISTOR 21.5K +-1% .125W TKF TC=0+-100 RESISTOR 215 +-1% .1W TKF TC=0+-100 RESISTOR 21.5K +-1% .125W TKF TC=0+-100 RESISTOR 10K +-1% .125W TKF TC=0+-100 RESISTOR 10K +-1% .125W TKF TC=0+-100 RESISTOR 3.16K +-1% .125W TKF TC=0+-100 RESISTOR 100K +-1% .125W TKF TC=0+-100 RESISTOR 1K +-1% .125W TKF TC=0+-100 RESISTOR 1.47K +-1% .125W TKF TC=0+-100 RESISTOR 8.25K +-1% .125W TKF TC=0+-100 RESISTOR 1K +-1% .125W TKF TC=0+-100 RESISTOR 46.4 +-1% .125W TKF TC=0+-100 RESISTOR .05 +-100% TKF RESISTOR 42.2K +-1% .125W TKF TC=0+-100 RESISTOR 14.7K +-1% .125W TKF TC=0+-100 RESISTOR 36.5K +-1% .125W TKF TC=0+-100 RESISTOR 215 +-1% .1W TKF TC=0+-100 RESISTOR 215 +-1% .1W TKF TC=0+-100 VOLTAGE SUPPRESSOR VMAC=230V, VMDC=300V SCREW- X 8MM-LG -HD SHIELD-DMM INTEGRATED CIRCUIT; ASCI CUST R NET PKG, REPLACES 34401-67901 IC OP AMP LP DUAL 8 PIN PLSTC-SOIC TRANSISTOR, JFET DUAL IC OP AMP WB SINGLE 8 PIN PLSTC-SOIC IC OP AMP LOW-NOISE SINGLE 8 PIN IC OP AMP LOW-BIAS-H-IMPD SINGLE 8 PIN IC Gate-ARY CMOS IC OP AMP LP DUAL 8 PIN PLSTC-SOIC
1 1 1 1 1 1 1
1 1 1 1 1 5 1 3 2 1 1
Mfr Code 04935 04935 04935 04935 04935 04935 04935 04935 04935 06337 04935 04935 04935 04935 04935 04935 04935 06337 04935 04935 02995 04935 04935 05524 06337 06337 06121
02361 02361 02632 02632 03285 03406 02037 03285 01698 03406 03285
194
Reference Designator U201 U301 U302 U303 U304 U305 U306 U307 U308 U309 U310
U311 U312 U400-U401 U402 U403 U411 U420 U450 XU403
Part Number 1826-2420 1826-2436 1826-2339 1826-4084 1826-1985 1826-4084 1826-1609 1827-0267 1826-2445 1820-5790 1826-1572
1820-5790 1826-4084 1826-2420 1826-1991 1826-1249 1821-3334 1826-1925 1818-6821 1200-1672
Qty
1 1 3 1 1 1 1 2 1
Part Description
IC OP AMP LP DUAL 8 PIN PLSTC-SOIC IC OP AMP WB 8 PIN PLSTC-SOIC IC; 8-BIT 16-P-SOIC CMOS IC OP AMP ANLG SINGLE 8 PIN PLSTC-SOIC ANALOG SWITCH 4 SPST 16 -P-SOIC IC OP AMP ANLG SINGLE 8 PIN PLSTC-SOIC ANALOG SWITCH 4 SPST 16 -P-SOIC MC34081BD RMS/DC 16-P-SOIC MISC IC SHF-RGTR CMOS/HC SYNCHRO SERIAL-IN IC COMPARATOR PRCN DUAL 8 PIN PLSTC-SOIC IC SHF-RGTR CMOS/HC SYNCHRO SERIAL-IN IC OP AMP ANLG SINGLE 8 PIN PLSTC-SOIC IC OP AMP LP DUAL 8 PIN PLSTC-SOIC IC OP AMP HS SINGLE 8 PIN PLSTC-SOIC IC, V RGLTR-V-REF-FXD IC-AN-MUX 74HC4053D 3X2:1 SOURCE-RESTRIC IC OP AMP LOW-NOISE SINGLE 8 PIN FRAM SERIAL 4K FMZ4C04-S SOCKET-IC-DIP 4-CONT DIP-SLDR
Mfr Code 03285 03406 03285 03285 02883 03285 03285 01698 03285 02910 02910
02910 03285 03285 03285 10858 02910 03285 14543 02194
1 1 1 1 1
195
Part Number 0160-7798 0160-7708 0160-7798 0160-7828 0160-5947 0160-7845 0160-7845 0160-5945 0160-7798 0180-4545 0160-5945 0160-5967 0160-5945 0160-5945 1906-0395 1902-1574 1902-1574 1906-0291 1906-0291 1906-0395 1906-0291 1906-0395 1902-1574
2110-0043 2110-0726 2110-0726 2110-0726 2110-0726 1252-8024 0360-2624 0360-2623 0360-2624 0490-1896 0490-1895
Qty
8 1 1 6 2 16
Part Description
CAP 0.1 uF 50V 10% X7R 0805 CAP-FXD 1000pF +-5% 50 V CER C0G CAP 0.1 uF 50V 10% X7R 0805 CAP .1 uF 16V 10% X7R CAP-FXD 1000pF 50 V CAP-FXD 180PF +-5% 50 V CER COG CAP-FXD 180PF +-5% 50 V CER COG CAP-FXD 0.01 uF 50 V CAP 0.1 uF 50V 10% X7R 0805 CAP-FXD 4.7 uF +-20% 20 V TA CAP-FXD 0.01 uF 50 V CAP-FXD 100pF +-5pct 50 V CER C0G CAP-FXD 0.01 uF 50 V CAP-FXD 0.01 uF 50 V DIODE-DUAL 75V TO-253 DIODE-ZNR 6.8V 6% TO-236 (SOT-23) DIODE-ZNR 6.8V 6% TO-236 (SOT-23) DIODE-DUAL 70V 100MA T0-236AA DIODE-DUAL 70V 100MA T0-236AA DIODE-DUAL 75V TO-253 DIODE-DUAL 70V 100MA T0-236AA DIODE-DUAL 75V TO-253 DIODE-ZNR 6.8V 6% TO-236 (SOT-23) FUSE (INCH) 1.5A 250V NTD FE UL-LST FUSEHOLDER-CLP-TYP FUSEHOLDER-CLP-TYP FUSEHOLDER-CLP-TYP FUSEHOLDER-CLP-TYP CONN-POST TYPE .100-PIN-SPCG-MTG-END CONN TERMINAL BK RA FEM 12PIN CONN TERMNAL BK RA FEM 4 PIN CONN TERMINAL BK RA FEM 12PIN RLY-2C 2A 220V 60WATT 3V-COIL LATCHING RLY-2C 2A 220V 60WATT 4.5V COIL
5 3 24
Mfr Code 02010 12340 02010 02010 02010 12340 12340 02010 02010 12340 02010 02010 02010 02010 02910 02910 02910 02910 02910 02910 02910 02910 02910
04703 02603 02603 02603 02603 05525 13389 13389 13389 12921 12921
2 4
1 4 1
24 3
196
Qty
Part Description
RLY-2C 2A 220V 60WATT 4.5V COIL RLY-2C 2A 220V 60WATT 4.5V COIL RLY-2C 2A 220V 60WATT 3V-COIL LATCHING
3 12
CORE-SHIELDING BEAD CORE-SHLD BEAD CORE-SHLD BEAD CORE-SHLD BEAD CORE-SHLD BEAD INDUCTOR 10UH +10% -10% 2.7W-MMX3.4LG-MM KIT, MODULE COVER, CASE
1 25 22
TRANSISTOR TRANSISTOR NPN SI SOT-23 (TO-236AB) TRANSISTOR PNP SI TO-236AA PD=200MW TRANSISTOR NPN SI SOT-23 (TO-236AB) TRANSISTOR PNP SI TO-236AA PD=200MW TRANSISTOR NPN SI SOT-23 (TO-236AB) TRANSISTOR PNP SI TO-236AA PD=200MW TRANSISTOR NPN SI SOT-23 (TO-236AB) TRANSISTOR PNP SI TO-236AA PD=200MW TRANSISTOR NPN SI SOT-23 (TO-236AB) TRANSISTOR PNP SI TO-236AA PD=200MW TRANSISTOR NPN SI SOT-23 (TO-236AB) TRANSISTOR PNP SI TO-236AA PD=200MW TRANSISTOR NPN SI SOT-23 (TO-236AB) TRANSISTOR PNP SI TO-236AA PD=200MW TRANSISTOR NPN SI SOT-23 (TO-236AB) TRANSISTOR PNP SI TO-236AA PD=200MW TRANSISTOR NPN SI SOT-23 (TO-236AB) TRANSISTOR PNP SI TO-236AA PD=200MW TRANSISTOR NPN SI SOT-23 (TO-236AB) TRANSISTOR PNP SI TO-236AA PD=200MW TRANSISTOR NPN SI SOT-23 (TO-236AB) TRANSISTOR PNP SI TO-236AA PD=200MW TRANSISTOR NPN SI SOT-23 (TO-236AB) TRANSISTOR PNP SI TO-236AA PD=200MW TRANSISTOR NPN SI SOT-23 (TO-236AB) TRANSISTOR PNP SI TO-236AA PD=200MW
197
Reference Designator Q309 Q310 Q311 Q312 Q313 Q314 Q315 Q316 Q317 Q318 Q319
R102-R103 R104 R105 R106-R109 R110-R112 R113 R114-R115 R117-R121 R147 R148 R149 R150-R154 R155 R158 R166-R168 R170-R180 R181 R182 R185 R201-R228 R301-R319 U101 U102-U106 U109
Part Number 1854-1053 1853-0525 1854-1053 1853-0525 1854-1053 1853-0525 1854-1053 1853-0525 1854-1053 1853-0525 1854-1053
0699-3034 0699-3947 0699-3034 0699-3970 0699-3974 0699-3970 0699-3067 0699-2973 0699-3051 0699-3077 0699-3051 0699-3034 0699-3947 0699-3947 0699-3034 0699-3963 0699-3947 0699-3932 0699-3932 0699-3051 0699-3051
Qty
Part Description
TRANSISTOR NPN SI SOT-23 (TO-236AB) TRANSISTOR PNP SI TO-236AA PD=200MW TRANSISTOR NPN SI SOT-23 (TO-236AB) TRANSISTOR PNP SI TO-236AA PD=200MW TRANSISTOR NPN SI SOT-23 (TO-236AB) TRANSISTOR PNP SI TO-236AA PD=200MW TRANSISTOR NPN SI SOT-23 (TO-236AB) TRANSISTOR PNP SI TO-236AA PD=200MW TRANSISTOR NPN SI SOT-23 (TO-236AB) TRANSISTOR PNP SI TO-236AA PD=200MW TRANSISTOR NPN SI SOT-23 (TO-236AB)
Mfr Code 12125 02237 12125 02237 12125 02237 12125 02237 12125 02237 12125
06337 06337 06337 06337 06337 06337 06337 06337 06337 06337 06337 06337 06337 06337 06337 06337 06337 06337 06337 06337 06337 01542 02910 03406
11 4 5 3 2 5 49 1
11 2
RESISTOR 1K +-1% .1W TKF TC=0+-100 RESISTOR 1K +-1% .063W TKF TC=0+-200 RESISTOR 1K +-1% .1W TKF TC=0+-100 RESISTOR 10K +-1% .063W TKF TC=0+-200 Resistor 14.7K 1% 0603 50V .063W TC=200 RES 10K 1% .063W RESISTOR 14.7K +-1% .1W TKF TC=0+-100 RES 215, FIXED THIN FILM RESISTOR 10K +-1% .1W TKF TC=0+-100 RESISTOR 1M +-1% .1W TKF TC=0+-100 RESISTOR 10K +-1% .1W TKF TC=0+-100 RESISTOR 1K +-1% .1W TKF TC=0+-100 RESISTOR 1K +-1% .063W TKF TC=0+-200 RESISTOR 1K +-1% .063W TKF TC=0+-200 RESISTOR 1K +-1% .1W TKF TC=0+-100 RESISTOR 4.64K 1% 0603 50V .063W TC=200 RESISTOR 1K +-1% .063W TKF TC=0+-200 RESISTOR 215 +-1% .063W TKF TC=0+-200 RESISTOR 215 +-1% .063W TKF TC=0+-200 RESISTOR 10K +-1% .1W TKF TC=0+-100 RESISTOR 10K +-1% .1W TKF TC=0+-100 PRGMD 1821-1876 IC FF CMOS/HC D-TYPE POS-EDGE-TRIG COM IC DCDR-DEMUXR CMOS/ACT BIN 8-TO-1-LINE
198
Qty
1 1 2 1 1
Part Description
IC SCHMITT-TRIG CMOS/ACT NAND QUAD 2-INP FRAM SERIAL 4K FMZ4C04-S IC INTERFACE MISC SOCKET-IC-PLCC 44-CONT SQUARE J-LEAD CERO-RES 12MHZ +1-0.8%
199
Part Number 0160-7798 0160-5967 0160-7828 0160-5947 0160-7845 0160-7845 0160-5945 0180-4545 0160-5945 0160-7708 0160-5945
1906-0395 1902-1574 1902-1574 1906-0291 1906-0395 1902-1572 1252-8024 0360-2624 0490-1897 0490-1897 9170-1663 9170-1584 9170-1584 9170-1584 9170-1584 9170-1663 9140-1638 34901-60001
Qty
4 8 1 1 2 7 1 3
Part Description
CAP 0.1UF 50V 10% X7R 0805 CAP-FXD 100PF 5% 0805 CAP .1UF 16V 10% X7R CAP-FXD 1000pF 50 V CAP FXD 180PF +-5% 50 V CER COG CAP FXD 180PF +-5% 50 V CER COG CAP-FXD 0.01uF 50 V CAP-FXD 4.7uF +-20% 20 V TA CAP-FXD 0.01uF 50 V CAP-FXD 1000pF +-5% 50 V CER C0G CAP-FXD 0.01uF 50 V DIODE-DUAL 75V TO-253 DIODE-ZNR 6.8V 6% TO-236 (SOT-23) DIODE-ZNR 6.8V 6% TO-236 (SOT-23) DIODE-DUAL 70V 100MA T0-236AA DIODE-DUAL 75V TO-253 DIODE-ZNR 15V 5% TO-236 (SOT-23) PD=.35W CONN-POST TYPE .100-PIN-SPCG-MTG-END CONN TERMINAL BK RA FEM 12PIN RELAY-REED 2A 1A 300VDC 5VDC-COIL RELAY-REED 2A 1A 300VDC 5VDC-COIL CORE-SHIELDING BEAD CORE-SHLD BEAD CORE-SHLD BEAD CORE-SHLD BEAD CORE-SHLD BEAD CORE-SHIELDING BEAD INDUCTOR 10UH +10% -10% 2.7W-MMX3.4LG-MM KIT, MODULE COVER, CASE
Mfr Code 02010 03292 02010 02010 12340 12340 02010 12340 02010 12340 02010
02910 02910 02910 02910 02910 02910 05525 13389 04501 04501 11702 09808 09808 09808 09808 11702 06352 02362
11 2 1 1 1 3 19
3 10
200
Qty
1 19
Part Description
TRANSISTOR TRANSISTOR NPN SI SOT-23 (TO-236AB) TRANSISTOR NPN SI SOT-23 (TO-236AB) RESISTOR 1K +-1% .1W TKF TC=0+-100 RESISTOR 1K +-1% .1W TKF TC=0+-100 RES 10K 1% .063W Resistor 14.7K 1% 0603 50V .063W TC=200 RES 10K 1% .063W RES 215, FIXED THIN FILM RESISTOR 14.7K +-1% .1W TKF TC=0+-100 RESISTOR 10K +-1% .1W TKF TC=0+-100 RESISTOR 1K +-1% .063W TKF TC=0+-200 RESISTOR 10K +-1% .1W TKF TC=0+-100 RESISTOR 1K +-1% .063W TKF TC=0+-200 RESISTOR 1M +-1% .1W TKF TC=0+-100 RESISTOR 10K +-1% .1W TKF TC=0+-100 RES 215, FIXED THIN FILM RESISTOR 1K +-1% .063W TKF TC=0+-200 RESISTOR 1K +-1% .063W TKF TC=0+-200 RES 215, FIXED THIN FILM RESISTOR 1K +-1% .1W TKF TC=0+-100 RESISTOR 4.64K 1% 0603 50V .063W TC=200 RESISTOR 215 +-1% .063W TKF TC=0+-200 RESISTOR 215 +-1% .063W TKF TC=0+-200 RESISTOR 10 1% 12181.0w tc=0+100 PRGMD 1821-1876 IC FF CMOS/HC D-TYPE POS-EDGE-TRIG COM IC SCHMITT-TRIG CMOS/ACT NAND QUAD 2-INP FRAM SERIAL 4K FMZ4C04-S IC INTERFACE MISC SOCKET-IC-PLCC 44-CONT SQUARE J-LEAD CERO-RES 12MHZ +1-0.8%
6 5 3 6 1 21 8
8 2 4 1 3 1 1 2 1 1
201
Part Number 0160-7798 0160-5947 0160-7798 0160-5947 0160-5967 0160-5945 0180-4545 0160-5945
1906-0395 1902-1544 1906-0291 1906-0395 1906-0395 1252-8024 0360-2624 0490-1896 0490-1896 9170-1663 9170-1584 9170-1584 34901-60001 1855-1101 1854-1053 1853-0525 1854-1053 1853-0525 1854-1053 1853-0525 1854-1053 1853-0525 1854-1053 1853-0525
Qty
5 4
Part Description
CAP 0.1UF +-10% 50V CER X7R CAP-FXD 1000pF 50 V CAP 0.1UF +-10% 50V CER X7R CAP-FXD 1000pF 50 V CAP-FXD 100pF +-5% 50 V CER C0G CAP-FXD 0.01uF 50 V CAP-FXD 4.7uF +-20% 20 V TA CAP-FXD 0.01uF 50 V DIODE-DUAL 75V TO-253 DIODE-ZNR 10V 5% TO-236 (SOT-23) PD=.35W DIODE-DUAL 70V 100MA T0-236AA DIODE-DUAL 75V TO-253 DIODE-DUAL 75V TO-253 CONN-POST TYPE .100-PIN-SPCG-MTG-END CONN TERMINAL BK RA FEM 12PIN RLY-2C 2A 220V 60WATT 3V-COIL LATCHING RLY-2C 2A 220V 60WATT 3V-COIL LATCHING CORE-SHIELDING BEAD CORE-SHLD BEAD CORE-SHLD BEAD KIT, MODULE COVER, CASE
8 9 1
Mfr Code 06352 02010 06352 02010 06352 02010 12340 02010
02910 02910 02910 02910 02910 05525 13389 12921 12921 11702 09808 09808 02362 02037 12125 02237 12125 02237 12125 02237 12125 02237 12125 02237
5 1 20
1 5 20
2 4
1 14 14
TRANSISTOR-MOSFET DUAL P-CHAN E-MODE SI TRANSISTOR NPN SI SOT-23 (TO-236AB) TRANSISTOR PNP SI TO-236AA PD=200MW TRANSISTOR NPN SI SOT-23 (TO-236AB) TRANSISTOR PNP SI TO-236AA PD=200MW TRANSISTOR NPN SI SOT-23 (TO-236AB) TRANSISTOR PNP SI TO-236AA PD=200MW TRANSISTOR NPN SI SOT-23 (TO-236AB) TRANSISTOR PNP SI TO-236AA PD=200MW TRANSISTOR NPN SI SOT-23 (TO-236AB) TRANSISTOR PNP SI TO-236AA PD=200MW
202
Reference Designator Q215 Q216-Q217 Q218 Q219 Q220-Q221 Q222-Q223 Q224 Q225 Q226-Q227 Q228
R102-R103 R105 R106-R109 R110-R112 R113 R114 R115 R116-R118 R120-R121 R122-R127 R140 R148 R149 R166-R168 R170-R177 R201-R228 U101 U102-U104 U141 U150 XU101 Y101
Part Number 1854-1053 1853-0525 1854-1053 1853-0525 1854-1053 1853-0525 1854-1053 1853-0525 1854-1053 1853-0525
0699-3034 0699-3034 0699-3051 0699-3067 0699-3051 0699-2973 0699-3067 0699-2973 0699-2973 0699-3034 0699-3051 0699-3077 0699-3051 0699-3034 0699-3044 0699-3051 34903-88801 1820-5752 1821-0055 1818-6821 1200-1592 0410-4009
Qty
Part Description
TRANSISTOR NPN SI SOT-23 (TO-236AB) TRANSISTOR PNP SI TO-236AA PD=200MW TRANSISTOR NPN SI SOT-23 (TO-236AB) TRANSISTOR PNP SI TO-236AA PD=200MW TRANSISTOR NPN SI SOT-23 (TO-236AB) TRANSISTOR PNP SI TO-236AA PD=200MW TRANSISTOR NPN SI SOT-23 (TO-236AB) TRANSISTOR PNP SI TO-236AA PD=200MW TRANSISTOR NPN SI SOT-23 (TO-236AB) TRANSISTOR PNP SI TO-236AA PD=200MW
Mfr Code 12125 02237 12125 02237 12125 02237 12125 02237 12125 02237
00746 00746 06337 00746 06337 06337 00746 06337 06337 00746 06337 00746 06337 00746 06337 06337 02632 02910 02037 14543 01380 00830
12 35 4 6
RESISTOR 1K +-1% .1W TKF TC=0+-100 RESISTOR 1K +-1% .1W TKF TC=0+-100 RESISTOR 10K +-1% .1W TKF TC=0+-100 RESISTOR 14.7K +-1% .1W TKF TC=0+-100 RESISTOR 10K +-1% .1W TKF TC=0+-100 RESISTOR 215 +-1% .1W TKF TC=0+-100 RESISTOR 14.7K +-1% .1W TKF TC=0+-100 RESISTOR 215 +-1% .1W TKF TC=0+-100 RESISTOR 215 +-1% .1W TKF TC=0+-100 RESISTOR 1K +-1% .1W TKF TC=0+-100 RESISTOR 10K +-1% .1W TKF TC=0+-100 RESISTOR 1M +-1% .1W TKF TC=0+-100 RESISTOR 10K +-1% .1W TKF TC=0+-100 RESISTOR 1K +-1% .1W TKF TC=0+-100 RESISTOR 4.64K +-1% .1W TKF TC=0+-100 RESISTOR 10K +-1% .1W TKF TC=0+-100 PROGMD 1821-1876 MCU N87C52 IC FF CMOS/HC D-TYPE POS-EDGE-TRIG COM IC SCHMITT-TRIG CMOS/ACT NAND QUAD 2-INP FRAM SERIAL 4K FMZ4C04-S SOCKET-IC-PLCC 44-CONT SQUARE J-LEAD CERO-RES 12MHZ +1-0.8%
1 3 1 1 1 1
203
Qty
1 1 12 1 2
Part Description
CAP 0.1UF 50V 10% X7R 0805 CAP-FXD 1000pF 50 V CAP-FXD 0.01uF 50 V CAP-FXD 4.7uF +-20% 20 V TA CF 100PF 5% 0805 CAP-FXD 0.01uF 50 V CAP-FXD 0.01uF 50 V DIODE-DUAL 70V 100MA T0-236AA DIODE-DUAL 75V TO-253 DIODE-DUAL 70V 100MA T0-236AA DIODE-DUAL 70V 100MA T0-236AA DIODE-DUAL 70V 100MA T0-236AA DIODE-DUAL 75V TO-253 CONN-POST TYPE .100-PIN-SPCG-MTG-END CONN TERMNAL BK RA FEM 4 PIN RLY-2C 2A 220V 60WATT 3V-COIL LATCHING RLY-2C 2A 220V 60WATT 3V-COIL LATCHING RLY-2C 2A 220V 60WATT 3V-COIL LATCHING RLY-2C 2A 220V 60WATT 3V-COIL LATCHING INDUCTOR 4.7uH +10% -10% CORE-SHLD BEAD KIT, MODULE COVER, CASE
33 3
1 6 32
2 5
1 12 12
TRANSISTOR TRANSISTOR PNP SI TO-236AA PD=200MW TRANSISTOR NPN SI SOT-23 (TO-236AB) TRANSISTOR PNP SI TO-236AA PD=200MW TRANSISTOR NPN SI SOT-23 (TO-236AB) TRANSISTOR PNP SI TO-236AA PD=200MW TRANSISTOR NPN SI SOT-23 (TO-236AB) TRANSISTOR PNP SI TO-236AA PD=200MW TRANSISTOR NPN SI SOT-23 (TO-236AB) TRANSISTOR PNP SI TO-236AA PD=200MW
204
Reference Designator Q213 Q221 Q223 Q231 Q233 Q241 Q243 Q251 Q253 Q261 Q263 Q271 Q273 Q281 Q283
R111 R113 R121 R123 R131 R133 R141 R143 R147 R148 R149 R151-R152 R154 R155 R156-R158 R160-R163 R164-R165 R166-R168 R211 R213 R221 R223 R231 R233
Part Number 1854-1053 1853-0525 1854-1053 1853-0525 1854-1053 1853-0525 1854-1053 1853-0525 1854-1053 1853-0525 1854-1053 1853-0525 1854-1053 1853-0525 1854-1053
0699-3051 0699-3051 0699-3051 0699-3051 0699-3051 0699-3051 0699-3051 0699-3051 0699-3051 0699-3053 0699-3051 0699-3034 0699-3034 0699-3061 0699-3067 0699-3051 0699-2973 0699-3034 0699-3051 0699-3051 0699-3051 0699-3051 0699-3051 0699-3051
Qty
Part Description
TRANSISTOR NPN SI SOT-23 (TO-236AB) TRANSISTOR PNP SI TO-236AA PD=200MW TRANSISTOR NPN SI SOT-23 (TO-236AB) TRANSISTOR PNP SI TO-236AA PD=200MW TRANSISTOR NPN SI SOT-23 (TO-236AB) TRANSISTOR PNP SI TO-236AA PD=200MW TRANSISTOR NPN SI SOT-23 (TO-236AB) TRANSISTOR PNP SI TO-236AA PD=200MW TRANSISTOR NPN SI SOT-23 (TO-236AB) TRANSISTOR PNP SI TO-236AA PD=200MW TRANSISTOR NPN SI SOT-23 (TO-236AB) TRANSISTOR PNP SI TO-236AA PD=200MW TRANSISTOR NPN SI SOT-23 (TO-236AB) TRANSISTOR PNP SI TO-236AA PD=200MW TRANSISTOR NPN SI SOT-23 (TO-236AB)
Mfr Code 12125 02237 12125 02237 12125 02237 12125 02237 12125 02237 12125 02237 12125 02237 12125
06337 06337 06337 06337 06337 06337 06337 06337 06337 02995 06337 06337 06337 06337 06337 06337 06337 06337 06337 06337 06337 06337 06337 06337
32
1 6 1 3 2
RESISTOR 10K +-1% .1W TKF TC=0+-100 RESISTOR 10K +-1% .1W TKF TC=0+-100 RESISTOR 10K +-1% .1W TKF TC=0+-100 RESISTOR 10K +-1% .1W TKF TC=0+-100 RESISTOR 10K +-1% .1W TKF TC=0+-100 RESISTOR 10K +-1% .1W TKF TC=0+-100 RESISTOR 10K +-1% .1W TKF TC=0+-100 RESISTOR 10K +-1% .1W TKF TC=0+-100 RESISTOR 10K +-1% .1W TKF TC=0+-100 RESISTOR 100K +-1% .1W TKF TC=0+-100 RESISTOR 10K +-1% .1W TKF TC=0+-100 RESISTOR 1K +-1% .1W TKF TC=0+-100 RESISTOR 1K +-1% .1W TKF TC=0+-100 RESISTOR 261 +-1% .1W TKF TC=0+-100 RESISTOR 14.7K +-1% .1W TKF TC=0+-100 RESISTOR 10K +-1% .1W TKF TC=0+-100 RES 215, FIXED THIN FILM RESISTOR 1K +-1% .1W TKF TC=0+-100 RESISTOR 10K +-1% .1W TKF TC=0+-100 RESISTOR 10K +-1% .1W TKF TC=0+-100 RESISTOR 10K +-1% .1W TKF TC=0+-100 RESISTOR 10K +-1% .1W TKF TC=0+-100 RESISTOR 10K +-1% .1W TKF TC=0+-100 RESISTOR 10K +-1% .1W TKF TC=0+-100
205
Reference Designator R241 R243 R251 R253 R261 R263 R271 R273 R281 R283-R284 R287
U101 U102-U103 U111 U141 XU101 Y101
Part Number 0699-3051 0699-3051 0699-3051 0699-3051 0699-3051 0699-3051 0699-3051 0699-3051 0699-3051 0699-3051 0699-3051
34903-88801 1820-5752 1818-6821 1821-0055 1200-1592 0410-4009
Qty
Part Description
RESISTOR 10K +-1% .1W TKF TC=0+-100 RESISTOR 10K +-1% .1W TKF TC=0+-100 RESISTOR 10K +-1% .1W TKF TC=0+-100 RESISTOR 10K +-1% .1W TKF TC=0+-100 RESISTOR 10K +-1% .1W TKF TC=0+-100 RESISTOR 10K +-1% .1W TKF TC=0+-100 RESISTOR 10K +-1% .1W TKF TC=0+-100 RESISTOR 10K +-1% .1W TKF TC=0+-100 RESISTOR 10K +-1% .1W TKF TC=0+-100 RESISTOR 10K +-1% .1W TKF TC=0+-100 RESISTOR 10K +-1% .1W TKF TC=0+-100
Mfr Code 06337 06337 06337 06337 06337 06337 06337 06337 06337 06337 06337
02632 02910 14543 02037 01380 00830
1 2 1 1 1 1
PROGMD 1821-1876 MCU N87C52 IC FF CMOS/HC D-TYPE POS-EDGE-TRIG COM FRAM SERIAL 4K FMZ4C04-S IC SCHMITT-TRIG CMOS/ACT NAND QUAD 2-INP SOCKET-IC-PLCC 44-CONT SQUARE J-LEAD CERO-RES 12MHZ +1-0.8%
206
34905A/34906A RF Multiplexer
Reference Designator C101-C103 C111 C140 C151 C158 C164-C165 C166 C201-C204 CR102 CR170
J101 K101-K103 K201-K203 K101-K103 K201-K203 L105-L109 L170-L171 MP1 P1-P10 P1-P10 Q101 Q102 Q103 Q104 Q105 Q106 Q107 Q108 Q109 Q110 Q111 Q112 Q170
Part Number 0160-5945 0160-5945 0160-7798 0160-5947 0180-4545 0160-5967 0160-5945 0160-5945 1906-0291 1906-0395
1252-8024 0490-1671 0490-1671 0490-1672 0490-1672 9170-1584 9140-1782
Qty
9 1 1 1 2
Part Description
CAP-FXD 0.01uF 50 V CAP-FXD 0.01uF 50 V CAP 0.1UF 50V 10% X7R 0805 CAP-FXD 1000pF 50 V CAP-FXD 4.7uF +-20% 20 V TA CF 100PF 5% 0805 CAP-FXD 0.01uF 50 V CAP-FXD 0.01uF 50 V DIODE-DUAL 70V 100MA T0-236AA DIODE-DUAL 75V TO-253 CONN-POST TYPE .100-PIN-SPCG-MTG-END RELAY, RG1ET-L-3V RELAY, RG1ET-L-3V For 34905A 50 RF Multiplexer: RELAY, 1C 3VDC-COIL 1A 24 VDC RELAY, 1C 3VDC-COIL 1A 24 VDC CORE-SHLD BEAD INDUCTOR 4.7uH +10% -10% KIT, MODULE COVER, CASE CONNECTOR-RF SMB JACK PC-W/O-STDF 50-OHM For 34906A 75 RF Multiplexer: CONNECTOR-RF SMB JACK PC-W/O-STDF 75-OHM TRANSISTOR NPN SI SOT-23 (TO-236AB) TRANSISTOR PNP SI TO-236AA PD=200MW TRANSISTOR NPN SI SOT-23 (TO-236AB) TRANSISTOR PNP SI TO-236AA PD=200MW TRANSISTOR NPN SI SOT-23 (TO-236AB) TRANSISTOR PNP SI TO-236AA PD=200MW TRANSISTOR NPN SI SOT-23 (TO-236AB) TRANSISTOR PNP SI TO-236AA PD=200MW TRANSISTOR NPN SI SOT-23 (TO-236AB) TRANSISTOR PNP SI TO-236AA PD=200MW TRANSISTOR NPN SI SOT-23 (TO-236AB) TRANSISTOR PNP SI TO-236AA PD=200MW TRANSISTOR-MOSFET DUAL P-CHAN E-MODE SI
1 1 1 6
Mfr Code 02010 02010 02010 02010 12340 03292 02010 02010 02910 02910
05525 01850 01850 01850 01850 09808 06352 02362 03621 03621 12125 02237 12125 02237 12125 02237 12125 02237 12125 02237 12125 02237 02037
6 5 2
34901-60001 1250-1377 10 1250-2339 1854-1053 1853-0525 1854-1053 1853-0525 1854-1053 1853-0525 1854-1053 1853-0525 1854-1053 1853-0525 1854-1053 1853-0525 1855-1101 10 12 36
207
Reference Designator Q201 Q202 Q203 Q204 Q205 Q206 Q207 Q208 Q209 Q210 Q211 Q212 Q231-Q254
R27 R101-R112 R140 R148 R149 R151-R152 R154 R155 R156-R158 R159-R163 R164-R165 R166-R168 R201-R212 SH101 SH103 SH201 SH203 U101 U102-U103 U111 U141 XU101 Y101
Part Number 1854-1053 1853-0525 1854-1053 1853-0525 1854-1053 1853-0525 1854-1053 1853-0525 1854-1053 1853-0525 1854-1053 1853-0525 1853-0525
0699-3034 0699-3040 0699-3051 0699-3053 0699-3051 0699-3034 0699-3034 0699-3061 0699-3067 0699-3051 0699-2973 0699-3034 0699-3040
Qty
Part Description
TRANSISTOR NPN SI SOT-23 (TO-236AB) TRANSISTOR PNP SI TO-236AA PD=200MW TRANSISTOR NPN SI SOT-23 (TO-236AB) TRANSISTOR PNP SI TO-236AA PD=200MW TRANSISTOR NPN SI SOT-23 (TO-236AB) TRANSISTOR PNP SI TO-236AA PD=200MW TRANSISTOR NPN SI SOT-23 (TO-236AB) TRANSISTOR PNP SI TO-236AA PD=200MW TRANSISTOR NPN SI SOT-23 (TO-236AB) TRANSISTOR PNP SI TO-236AA PD=200MW TRANSISTOR NPN SI SOT-23 (TO-236AB) TRANSISTOR PNP SI TO-236AA PD=200MW TRANSISTOR PNP SI TO-236AA PD=200MW RESISTOR 1K +-1% .1W TKF TC=0+-100 RESISTOR 3.16K +-1% .1W TKF TC=0+-100 RESISTOR 10K +-1% .1W TKF TC=0+-100 RESISTOR 100K +-1% .1W TKF TC=0+-100 RESISTOR 10K +-1% .1W TKF TC=0+-100 RESISTOR 1K +-1% .1W TKF TC=0+-100 RESISTOR 1K +-1% .1W TKF TC=0+-100 RESISTOR 261 +-1% .1W TKF TC=0+-100 RESISTOR 14.7K +-1% .1W TKF TC=0+-100 RESISTOR 10K +-1% .1W TKF TC=0+-100 RESISTOR 215 +-1% .1W TKF TC=0+-100 RESISTOR 1K +-1% .1W TKF TC=0+-100 RESISTOR 3.16K +-1% .1W TKF TC=0+-100 SHIELD-RF SHIELD-RF SHIELD-RF SHIELD-RF PROGMD 1821-1876 MCU N87C52 IC FF CMOS/HC D-TYPE POS-EDGE-TRIG COM FRAM SERIAL 4K FMZ4C04-S IC SCHMITT-TRIG CMOS/ACT NAND QUAD 2-INP SOCKET-IC-PLCC 44-CONT SQUARE J-LEAD CERO-RES 12MHZ +1-0.8%
Mfr Code 12125 02237 12125 02237 12125 02237 12125 02237 12125 02237 12125 02237 02237
06337 02995 06337 02995 06337 06337 06337 06337 06337 06337 06337 06337 02995 01642 01642 01642 01642 02632 02910 14543 02037 01380 00830
24 7 1 7 1 3 2
34905-60601 4 34905-60601 34905-60601 34905-60601 34903-88801 1820-5752 1818-6821 1821-0055 1200-1592 0410-4009 1 2 1 1 1 1
208
Part Description
CAP 0.1UF 50V 10% X7R 0805 CAP-FXD 0.01uF 50 V CAP 0.1UF 50V 10% X7R 0805 CAP-FXD 0.01uF 50 V CAP-FXD 100pF +-5% 50 V CER C0G CAP-FXD 1000pF 50 V CAP 0.1UF 50V 10% X7R 0805 CAP-FXD 4.7uF +-20% 20 V TA CAP-FXD 0.01uF 50 V CAP 0.1UF 50V 10% X7R 0805 CAP-FXD 0.01uF 50 V CAP 0.1UF 50V 10% X7R 0805 CAP-FXD 10uF +-20% 35 V TA CAP-FXD 100pF +-5% 50 V CER C0G CAP 0.1UF 50V 10% X7R 0805 CAP-FXD 0.01uF 50 V CAP 0.1UF 50V 10% X7R 0805 CAP-FXD 0.01uF 50 V CAP 0.1UF 50V 10% X7R 0805 CAP-FXD 0.01uF 50 V CAP 0.1UF 50V 10% X7R 0805 CAP-FXD 22uF 20 V TA CAP-FXD 4700pF 50 V CAP-FXD 0.047uF 50 V CAP-FXD 10uF +-20% 35 V TA CAP-FXD 1uF +-20% 35 V TA CAP-FXD 10uF +-20% 35 V TA TANT SMD 4.7 20% CAP 0.1UF 50V 10% X7R 0805 CAP-FXD 10uF +-20% 35 V TA CAP-FXD 100pF +-5% 50 V CER C0G CAP-FXD 1000pF 50 V CAP-FXD 100pF +-5% 50 V CER C0G CAP-FXD 1000pF 50 V CAP-FXD 100pF +-1% 50 V CER C0G CAP-FXD 1000pF 50 V
9 4 1
1 1 1 2 1 1
Mfr Code 02010 02010 02010 02010 03292 02010 02010 12340 02010 02010 02010 02010 05524 03292 02010 02010 02010 02010 02010 02010 02010 00039 12473 12473 05524 00039 05524 00039 02010 05524 03292 02010 03292 02010 12340 02010
209
Reference Designator CR101-CR106 CR110 CR301-CR308 CR401-CR408 CR501-CR504 CR505-CR506 CR507 CR510-CR511 CR520-CR521 CR530 CR531
FB501-FB506 J101 L105 L109 L201-L202 L501-L502 L510 L515-L516 MAL510 MAP102 MP1 P101 P102 P301 P401 P501 Q301-Q308 Q401-Q408 R101-R102 R104 R105
Part Number 1906-0358 1906-0291 1906-0358 1906-0358 1901-1386 1902-1643 1901-1582 1901-1332 1906-0291 1901-1348 1906-0291
9170-1584 1252-8024 9170-1584 9170-1584 9140-1782 9140-1638 9140-2152 9140-1238
Qty
22 4
Part Description
DIODE-DUAL 100V 100MA TO-236AB (SOT-23) DIODE-DUAL 70V 100MA T0-236AA DIODE-DUAL 100V 100MA TO-236AB (SOT-23) DIODE-DUAL 100V 100MA TO-236AB (SOT-23) DIO SI PN SOT23 100V 750MA BAS78B DIODE-ZNR 15V PD=1.5W IR=1UA DIODE-PWR RECT 30V 500MA DIODE-UNMOUNTED CHIP DIODE-DUAL 70V 100MA T0-236AA DIO TVS UNIDIR SMB 24V 600WP SMBJ24A DIODE-DUAL 70V 100MA T0-236AA CORE-SHLD BEAD CONN-POST TYPE .100-PIN-SPCG-MTG-END CORE-SHLD BEAD CORE-SHLD BEAD INDUCTOR 4.7uH +10% -10% INDUCTOR 10UH +10% -10% 2.7W-MMX3.4LG-MM L SMT 22uH 20% 1.32A INDUCTOR 10UH +-5% 2.8W-MMX3.4LG-MM Q=30 SHIELD JUMPER - REMOVABLE, TWO POSITION KIT, MODULE COVER, CASE CONN TERMNAL BK RA FEM 4 PIN CONN-POST TYPE .100-PIN-SPCG 3-CONT CONN TERMINAL BK RA FEM 12PIN CONN TERMINAL BK RA FEM 12PIN CONN TERMNAL BK RA FEM 4 PIN TRANSISTOR MOSFET N-CHAN E-MODE TO-252AA TRANSISTOR MOSFET N-CHAN E-MODE TO-252AA RESISTOR 1K +-1% .1W TKF TC=0+-100 RESISTOR 1K +-1% .1W TKF TC=0+-100 RESISTOR 100K +-1% .1W TKF TC=0+-100
4 2 1 2 1
Mfr Code 03406 02910 03406 03406 06121 02037 02037 03038 02910 03287 02910
09808 05525 09808 09808 06352 06352 05524 02366 02631 01380 02362 13389 04726 13389 13389 13389 02037 02037 06337 06337 02995
8 1
2 2 1 2
34907-00601 1 1258-0209 1 34901-60001 0360-2623 2 1251-4670 1 0360-2624 2 0360-2624 0360-2623 1855-0800 1855-0800 0699-3034 0699-3034 0699-3053 16
8 4
210
Reference Designator R108 R109 R110-R111 R112 R113 R114 R115 R116 R117 R118 R119 R120-R123 R124 R125-R128 R129 R130-R133 R134-R135 R140 R141 R148 R151-R153 R166-R168 R205 R211-R214 R215-R216 R301 R302 R501 R502 R503 R504 R505-R506 R510 R511 R512 R513 R514 R518 R531-R534 R550-R551
Part Number 0699-3053 0699-3051 0699-2973 0699-3053 0699-3070 0699-3044 0699-3052 0699-2997 0699-3052 0699-3044 0699-2986 0699-1319 0699-3008 0699-1319 0699-3008 0699-1319 0699-3008 0699-3052 0699-3049 0699-3053 0699-2643 0699-3034 0699-3051 0699-2973 0699-3051 0699-3047 0699-3045 0699-3602 0699-2838 0699-3602 0699-2838 0699-3052 0699-3036 0699-3046 0699-2962 0699-3044 0699-3034 0699-3034 0699-2973 0699-3832
Qty
4 10 1 3 5 1
Part Description
RESISTOR 100K +-1% .1W TKF TC=0+-100 RESISTOR 10K +-1% .1W TKF TC=0+-100 RESISTOR 215 +-1% .1W TKF TC=0+-100 RESISTOR 100K +-1% .1W TKF TC=0+-100 RESISTOR 26.1K +-1% .1W TKF TC=0+-100 RESISTOR 4.64K +-1% .1W TKF TC=0+-100 RESISTOR 12.1K +-1% .1W TKF TC=0+-100 RESISTOR 133K +-1% .1W TKF TC=0+-100 RESISTOR 12.1K +-1% .1W TKF TC=0+-100 RESISTOR 4.64K +-1% .1W TKF TC=0+-100 RESISTOR 21.5K +-1% .1W TKF TC=0+-100 RESISTOR 12.1K +-1% .125W TKF TC=0+-100 RESISTOR 511K +-1% .1W TKF TC=0+-100 RESISTOR 12.1K +-1% .125W TKF TC=0+-100 RESISTOR 511K +-1% .1W TKF TC=0+-100 RESISTOR 12.1K +-1% .125W TKF TC=0+-100 RESISTOR 511K +-1% .1W TKF TC=0+-100 RESISTOR 12.1K +-1% .1W TKF TC=0+-100 RESISTOR 8.25K +-1% .1W TKF TC=0+-100 RESISTOR 100K +-1% .1W TKF TC=0+-100 RESISTOR 0 +-5% .1W TKF TC=0+-300 RESISTOR 1K +-1% .1W TKF TC=0+-100 RESISTOR 10K +-1% .1W TKF TC=0+-100 RESISTOR 215 +-1% .1W TKF TC=0+-100 RESISTOR 10K +-1% .1W TKF TC=0+-100 RESISTOR 6.81K +-1% .1W TKF TC=0+-100 RESISTOR 5.11K +-1% .1W TKF TC=0+-100 RESISTOR 12.4K +-0.1% .125W TF TC=0+-25 RESISTOR 40K +-0.1% .125W TF TC=0+-25 RESISTOR 12.4K +-0.1% .125W TF TC=0+-25 RESISTOR 40K +-0.1% .125W TF TC=0+-25 RESISTOR 12.1K +-1% .1W TKF TC=0+-100 RESISTOR 1.96K +-1% .1W TKF TC=0+-100 RESISTOR 6.19K +-1% .1W TKF TC=0+-100 RESISTOR 68.1K +-1% .1W TKF TC=0+-100 RESISTOR 4.64K +-1% .1W TKF TC=0+-100 RESISTOR 1K +-1% .1W TKF TC=0+-100 RESISTOR 1K +-1% .1W TKF TC=0+-100 RESISTOR 215 +-1% .1W TKF TC=0+-100 RESISTOR 51.1 +-1% .1W TKF TC=0+-100
1 16 4
1 3
1 1 2 2
1 1 1
Mfr Code 02995 06337 06337 02995 06337 06337 02995 06337 02995 06337 06337 04935 06337 04935 06337 04935 06337 02995 06337 02995 06337 06337 06337 06337 06337 02995 06337 06337 06337 06337 06337 02995 06337 06337 06337 06337 06337 06337 06337 06337
211
Qty
1
Part Description
RESISTOR 261 +-1% .1W TKF TC=0+-100 RESISTOR 12.1K +-1% .125W TKF TC=0+-100 RESISTOR; CARBON FILM;NTWRK RESISTOR; CARBON FILM;NTWRK NETWORK-RES 620 X 4 5% 1206 .125W 200V NETWORK-RES 3.3KX4 5% 1206 .125W 200W RESISTOR; CARBON FILM;NTWRK NETWORK-RES 100K X 4 5% 1206 .125W 200V RESISTOR; CARBON FILM;NTWRK NETWORK-RES 3.3KX4 5% 1206 .125W 200W RESISTOR; CARBON FILM;NTWRK NETWORK-RES 100K X 4 5% 1206 .125W 200V RESISTOR; CARBON FILM;NTWRK PROGRAM PART FRAM SERIAL 4K FMZ4C04-S IC SCHMITT-TRIG CMOS/ACT INV HEX IC GATE CMOS/HC OR QUAD 2-INP IC FF CMOS/HC D-TYPE POS-EDGE-TRIG COM IC SCHMITT-TRIG CMOS/ACT NAND QUAD 2-INP IC OP AMP LP DUAL 8 PIN PLSTC-SOIC IC COMPARATOR GP SINGLE 8 PIN PLSTC-SOIC IC SHF-RGTR CMOS/HC SYNC/ASYNC IC GATE CMOS/HC EXCL-OR QUAD 2-INP IC FF CMOS/HC D-TYPE POS-EDGE-TRIG COM IC DRVR CMOS/HC BUS OCTL IC COMPARATOR LP QUAD 14 PIN PLSTC-SOIC IC COMPARATOR LP QUAD 14 PIN PLSTC-SOIC IC PWR MGT-V-REG-ADJ-NEG -5.25/-4.75V 8 D/A 16-BIT 16-P-SOIC BICMOS IC OP AMP PRCN SINGLE 8 PIN PLSTC-SOIC IC PWR MGT-V-REG-SWG 8 PINS P-SOIC PKG DIO TVS BIDIR SMB 600WP 110V SMBJ110C SOCKET-IC-PLCC 44-CONT SQUARE J-LEAD CERO-RES 12MHZ +1-0.8%
12 4 4 4
1 1 1 1 6 1 1 1 1 1 2 4 1 2 2 1 4 1 1
212
Part Number 0160-7798 0160-7708 0160-7798 0160-7828 0160-5947 0160-7845 0160-7845 0160-5945 0160-7798 0180-4545 0160-5945 0160-5945 0160-7798
1906-0395 1902-1574 1902-1574 1906-0291 1906-0291 1906-0395 1906-0291 1906-0395 1252-8024 0360-2624 0360-2624 0490-1896 0490-1895 0490-1896 9170-1663 9170-1584 9170-1584 9170-1584 9170-1584 9170-1663
Qty
9 1 1 5 2 12 1
Part Description
CAP 0.1UF 50V 10% X7R 0805 CAP-FXD 1000pF +-5% 50 V CER C0G CAP 0.1UF 50V 10% X7R 0805 CAP .1UF 16V 10% X7R CAP-FXD 1000pF 50 V CAP FXD 180PF +-5% 50 V CER COG CAP FXD 180PF +-5% 50 V CER COG CAP-FXD 0.01uF 50 V CAP 0.1UF 50V 10% X7R 0805 CAP-FXD 4.7uF +-20% 20 V TA CAP-FXD 0.01uF 50 V CAP-FXD 0.01uF 50 V CAP 0.1UF 50V 10% X7R 0805 DIODE-DUAL 75V TO-253 DIODE-ZNR 6.8V 6% TO-236 (SOT-23) DIODE-ZNR 6.8V 6% TO-236 (SOT-23) DIODE-DUAL 70V 100MA T0-236AA DIODE-DUAL 70V 100MA T0-236AA DIODE-DUAL 75V TO-253 DIODE-DUAL 70V 100MA T0-236AA DIODE-DUAL 75V TO-253 CONN-POST TYPE .100-PIN-SPCG-MTG-END CONN TERMINAL BK RA FEM 12PIN CONN TERMINAL BK RA FEM 12PIN RLY-2C 2A 220V 60WATT 3V-COIL LATCHING RLY-2C 2A 220V 60WATT 4.5V COIL RLY-2C 2A 220V 60WATT 3V-COIL LATCHING CORE-SHIELDING BEAD CORE-SHLD BEAD CORE-SHLD BEAD CORE-SHLD BEAD CORE-SHLD BEAD CORE-SHIELDING BEAD
Mfr Code 02010 12340 02010 02010 02010 12340 12340 02010 02010 12340 02010 02010 02010
02910 02910 02910 02910 02910 02910 02910 02910 05525 13389 13389 12921 12921 12921 11702 09808 09808 09808 09808 11702
5 2 22
1 4
21 1
4 10
213
Part Qty Part Description Number 34901-60001 KIT, MODULE COVER, CASE
1855-1101 1854-1053 1853-0525 1854-1053 1853-0525 1854-1053 1853-0525 1854-1053 1853-0525 1854-1053 1853-0525 1854-1053 1853-0525 1854-1053 1853-0525 1854-1053 1853-0525 1854-1053 1853-0525 1854-1053 1853-0525 1854-1053 1853-0525 1854-1053 1853-0525 1854-1053 0699-3034 0699-3051 0699-3067 0699-3051 0699-3067 0699-2973 0699-3051 0699-3077 0699-3051 0699-3034 1 17 16 TRANSISTOR-MOSFET DUAL P-CHAN E-MODE SI TRANSISTOR NPN SI SOT-23 (TO-236AB) TRANSISTOR PNP SI TO-236AA PD=200MW TRANSISTOR NPN SI SOT-23 (TO-236AB) TRANSISTOR PNP SI TO-236AA PD=200MW TRANSISTOR NPN SI SOT-23 (TO-236AB) TRANSISTOR PNP SI TO-236AA PD=200MW TRANSISTOR NPN SI SOT-23 (TO-236AB) TRANSISTOR PNP SI TO-236AA PD=200MW TRANSISTOR NPN SI SOT-23 (TO-236AB) TRANSISTOR PNP SI TO-236AA PD=200MW TRANSISTOR NPN SI SOT-23 (TO-236AB) TRANSISTOR PNP SI TO-236AA PD=200MW TRANSISTOR NPN SI SOT-23 (TO-236AB) TRANSISTOR PNP SI TO-236AA PD=200MW TRANSISTOR NPN SI SOT-23 (TO-236AB) TRANSISTOR PNP SI TO-236AA PD=200MW TRANSISTOR NPN SI SOT-23 (TO-236AB) TRANSISTOR PNP SI TO-236AA PD=200MW TRANSISTOR NPN SI SOT-23 (TO-236AB) TRANSISTOR PNP SI TO-236AA PD=200MW TRANSISTOR NPN SI SOT-23 (TO-236AB) TRANSISTOR PNP SI TO-236AA PD=200MW TRANSISTOR NPN SI SOT-23 (TO-236AB) TRANSISTOR PNP SI TO-236AA PD=200MW TRANSISTOR NPN SI SOT-23 (TO-236AB) RESISTOR 1K +-1% .1W TKF TC=0+-100 RESISTOR 10K +-1% .1W TKF TC=0+-100 RESISTOR 14.7K +-1% .1W TKF TC=0+-100 RESISTOR 10K +-1% .1W TKF TC=0+-100 RESISTOR 14.7K +-1% .1W TKF TC=0+-100 RESISTOR 215 +-1% .1W TKF TC=0+-100 RESISTOR 10K +-1% .1W TKF TC=0+-100 RESISTOR 1M +-1% .1W TKF TC=0+-100 RESISTOR 10K +-1% .1W TKF TC=0+-100 RESISTOR 1K +-1% .1W TKF TC=0+-100
12 40 5
4 1
214
Reference Designator R154 R156 R166-R168 R170-R180 R181 R182 R185 R201-R233 U101 U102-U105 U109 U141 U150 U151-U152
XU101 Y101
Part Number 0699-3947 0699-3947 0699-3034 0699-3044 0699-3034 0699-3932 0699-3932 0699-3051 34901-88822 1820-5752 1820-5941 1821-0055 1818-6821 1821-2382
1200-1592 0410-4009
Qty
2
Part Description
RESISTOR 1K +-1% .063W TKF TC=0+-200 RESISTOR 1K +-1% .063W TKF TC=0+-200 RESISTOR 1K +-1% .1W TKF TC=0+-100 RESISTOR 4.64K +-1% .1W TKF TC=0+-100 RESISTOR 1K +-1% .1W TKF TC=0+-100 RESISTOR 215 +-1% .063W TKF TC=0+-200 RESISTOR 215 +-1% .063W TKF TC=0+-200 RESISTOR 10K +-1% .1W TKF TC=0+-100 PRGMD 1821-1876 IC FF CMOS/HC D-TYPE POS-EDGE-TRIG COM IC DCDR-DEMUXR CMOS/ACT BIN 8-TO-1-LINE IC SCHMITT-TRIG CMOS/ACT NAND QUAD 2-INP FRAM SERIAL 4K FMZ4C04-S IC-INTERFACE MISC BIPOLAR 9-BIT SOCKET-IC-PLCC 44-CONT SQUARE J-LEAD CERO-RES 12MHZ +1-0.8%
11 2
1 4 1 1 1 2 1 1
Mfr Code 06337 06337 06337 06337 06337 06337 06337 06337 02632 02910 03406 02037 14543 12186
01380 00830
215
Manufacturers List
Mfr Code
00039 00746 00830 01380 01542 01642 01698 01850 02010 02037 02194 02237 02361/2 02499 02883 02910 03038 03285 03287 03292 03406 03418 03677 03744 03811 03827 04078 04501 04504 04670 04726 ROHM CORP KYOCERA AMERICA, INC AMP INC DIV 01 SAN JOSE COMPONENTS SONS TOOL INC TEXAS INSTRUMENTS INC AROMAT CORP AVX CORP MOTOROLA INC ROBINSON NUGENT INC FAIRCHILD SEMICONDUCTOR AGILENT TECHNOLOGIES INTERNATIONAL RESISTIVE CO. TEMIC/SILICONIX INC PHILIPS SEMICONDUCTORS INTL RECTIFIER CORP ANALOG DEVICES INC GENERAL SEMICONDUCTOR IND INC CORNING GLASS WORKS NATIONAL SEMICONDUCTOR CORP MOLEX INC AMERICAN MICRO SYSTEMS INC BOURNS NETWORKS INC INTEL CORP FAIR RITE PRODUCTS CORP SGS-THOMSON MICROELECTRONICS INC COTO WABASH GENERAL INSTRUMENT CORP JOHNSON COMPONENTS INC 3M CO
Mfr Name
NEC ELECTRONICS INC MTN VIEW
City
KYOTO 615 SAN DIEGO HARRISBURG SAN JOSE WOODVILLE DALLAS MOUNTAINSIDE GREAT NECK ROSELLE NEW ALBANY SOUTH PORTLAND LOVELAND BOONE SANTA CLARA EINDHOVEN LOS ANGELES NORWOOD TEMPE CORNING SANTA CLARA LISLE SANTA CLARA RIVERSIDE SANTA CLARA WALLKILL PHOENIX PROVIDENCE CHICAGO WASECA ST PAUL
State
CA CA PA CA WI TX NJ NY IL IN ME CO NC CA CA MA AZ NY CA IL CA CA CA NY AZ RI IL MN MN
Country
US JP US US US US US US US US US US US US US NL US US US US US US US US US US US US US US US
216
Mfr Code
04733 05176 05524 05525 05535 05951 06121 06337 06352 06360 06916 07179 07371 08709 09235 09454 09808 09891 09939 10421 10858 11302 11484 11702 11908 12125 12186 12340 12768 12921 13389 14543
Mfr Name
BELL INDUSTRIES INC MILLER JW DIV AMERICAN SHIZUKI CORP VISHAY INTERTECHNOLOGY INC ELCO CORP KEYSTONE ELECTRONICS CORP WICKMANN-WERKE A G SIEMENS AG PHILIPS ELECTRONICS NV TDK CORPORATION OF AMERICA NIPPON CHEMI-CON CORP SONY CORP AAVID THERMAL TECHNOLOGY INC SANYO ELECTRIC INC MATSUSHITA ELECTRIC CORPORATION OF FOX ELECTRONICS PRECISION RESISTIVE PRODS INC STEWARD INC KOA CORPORATION MURATA ELECTRONICS NORTH AMERICA, INC. EPSON AMERICA INC LINEAR TECHNOLOGY CORP MAXIM INTEGRATED PRODUCTS JOSLYN ELECTRONIC SYSTEMS DIV TAIYO YUDEN CO LTD NORITAKE CO LTD SAMSUNG SEMICONDUCTOR INC DALLAS SEMICONDUCTOR CORP KEMET ELECTRONICS CORPORATION SEIKO EPSON CORP OMRON JAPAN WIELAND INC RAMTRON GARDENA
City
CANOGA PARK MALVERN NEWPORT BEACH NEW YORK WITTEN-ANNEN MUNICH EINDHOVEN SKOKIE OHME-SHI TOKYO TOYKO LACONIA MORIGYCHI SECAUCUS FT MEYERS MEDIAPOLIS CHATTANOOGA INA-SHI NAGANO-KEN LONG BEACH TORRENCE MILPITAS SUNNYVALE GOLETA TOKYO 110 NISHI-KU - NAGOYA SANTA CLARA DALLAS GREENVILLE NAJANO-KEN TOKYO KYOTO ROCHELLE COLORADO SPRINGS
State
CA CA PA CA NY
Country
US US US US US DE DE NL
IL
US JP JP
NH NJ FL IA TN CA CA CA CA CA
US JP US US US US JP US US US US US JP JP
CA TX SC
US US US JP JP
NY CO
US US
217
218
Schematics
Schematics
This chapter contains a block diagram, schematics, and component locator drawings for the instrument and all plug-in modules. The block diagram and schematics support the Theory of Operation in chapter 5. Agilent 34970A System Block Diagram, o n page 221 A1 Component Locators, starting on page 222 A1 Schematics, starting on page 224 A2 Display and Keyboard Component Locator, on page 228 A2 Display and Keyboard Schematic, on page 229 A3 Backplane Component Locator, o n page 230 A3 Backplane Schematic, o n page 231 A4 Internal DMM Component Locators, starting o n page 232 A4 Internal DMM Schematics, starting o n page 234 34901A 20-Channel Multiplexer Component Locator, on page 238 34901A 20-Channel Multiplexer Schematics, starting on page 239 34902A 16-Channel Multiplexer Component Locator, on page 244 34902A 16-Channel Multiplexer Schematics, starting on page 245 34903A 20-Channel Actuator Component Locator, o n page 249 34903A 20-Channel Actuator Schematics, starting o n page 250 34904A 4x8 Matrix Component Locator, o n page 253 34904A 4x8 Matrix Schematics, starting o n page 254 34905A/34906A RF Multiplexer Component Locator, on page 257 34905A/34906A RF Multiplexer Schematics, starting o n page 258 34907A Multifunction Module Component Locator, on page 260 34907A Multifunction Module Schematics, starting on page 261 34908A 40-Channel Multiplexer Component Locator, on page 266 34908A 40-Channel Multiplexer Schematics, starting on page 267
220
Note: The DIAG:DMM:CYCLes? command returns three numbers indicating the cycle count on backplane relays 1, 2, and 3 (which correspond to relays K102, K103, and K104 respectively). See chapter 5 in the Users Guide for more information.
Specifications
The table below shows the state of the instrument after a FACTORY RESET from the Sto/Rcl menu or *RST command from the remote interface.
Quick Start
Measurement Configuration Function Range Resolution Integration Time Input Resistance Channel Delay Totalizer Reset Mode Totalizer Edge Detect Scanning Operations Scan List Reading Memory Min, Max, and Average Scan Interval Source Scan Interval Scan Count Scan Reading Format Monitor in Progress Factory Reset State DC Volts Autorange 512 digits 1 PLC 10 M (fixed for all DCV ranges) Automatic Delay Count Not Reset When Read Rising Edge Factory Reset State Empty All Readings are Cleared All Statistical Data is Cleared Immediate Front Panel = 10 Seconds Remote = Immediate Front Panel = Continuous Remote = 1 Scan Sweep Reading Only (No Units, Channel, Time) Stopped Factory Reset State 1 0 Vdc Factory Reset State Not Cleared Off 0 Alarm 1 Latched Mode Output Lines are Cleared Fail = Low Factory Reset State Reset: All Channels Open Reset: All Channels Open Reset: Channels s11 and s21 Selected Reset: Both DIO Ports = Input, Count = 0, Both DACs = 0 Vdc Factory Reset State On Errors Not Cleared No Change
Front-Panel Overview
Calibration Procedures
Theory of Operation
Mx+B Scaling Gain Factor (M) Scale Factor (B) Scale Label Alarm Limits Alarm Queue Alarm State HI and LO Alarm Limits Alarm Output Alarm Output Configuration Alarm Output State Alarm Output Slope Module Hardware 34901A, 34902A, 34908A 34903A, 34904A 34905A, 34906A 34907A System-Related Operations Display State Error Queue Stored States
Service
Replaceable Parts
Schematics