Location via proxy:   [ UP ]  
[Report a bug]   [Manage cookies]                

Accutorr V Service Manual

Download as pdf or txt
Download as pdf or txt
You are on page 1of 125

VAULT COPY

Accutorr is a registered trademark of Mindray DS USA, Inc. Masimo SET is a U.S. registered trademark of Masimo Corp. Mindray is a trademark or a registered trademark of Shenzhen Mindray Bio-Medical Electronics Co., Ltd. Nellcor is a U.S. registered trademark of Tyco International Ltd. SmarTemp is a trademark of Shenzhen Mindray Bio-Medical Electronics Co., Ltd.

Copyright Mindray DS USA, Inc., 2009. All rights reserved. Contents of this publication may not be reproduced in any form without permission of Mindray DS USA, Inc.

Accutorr V Service Manual

0070-10-0702

Table of Contents

Introduction ....................................................................................................................................................1 - 2 Controls and Indicators....................................................................................................................................1 - 2 System Overview ............................................................................................................................................1 - 2 Hardware Structure .........................................................................................................................................1 - 4 Main Board ............................................................................................................................................1 - 5 Power Board...........................................................................................................................................1 - 6 Key and Displays Board ...........................................................................................................................1 - 7 Parameter Boards ....................................................................................................................................1 - 8 Recorder ................................................................................................................................................1 - 10 Bar Code Scanner ...................................................................................................................................1 - 11 Introduction ....................................................................................................................................................2 - 2 NIBP Measurement .........................................................................................................................................2 - 2 Blood Pressure ........................................................................................................................................2 - 2 Static Accuracy .......................................................................................................................................2 - 2 Maximum Cuff Pressure ............................................................................................................................2 - 2 Cuff Inflation ...........................................................................................................................................2 - 3 Maximum Leakage ..................................................................................................................................2 - 3 Deflation Rate .........................................................................................................................................2 - 3 Update Period.........................................................................................................................................2 - 3 Initial NIBP Pressure Setup and Range........................................................................................................2 - 3 NIBP Measurement Cycle.................................................................................................................................2 - 4 Cuff Inflation Time....................................................................................................................................2 - 4 Automatically Adjusted Inflation Value after First Measurement......................................................................2 - 5 Pulse Rate ...............................................................................................................................................2 - 5 SpO2 Measurement ........................................................................................................................................2 - 6 DPM SpO2 Module Performance...............................................................................................................2 - 6 Masimo MS-2013 SpO2 Performance........................................................................................................2 - 7 Nellcor NELL-3 SpO2 Performance.............................................................................................................2 - 9 Temperature ...................................................................................................................................................2 - 10 Display Area, Indicator, and Controller .............................................................................................................2 - 10 Display Area...........................................................................................................................................2 - 10 Overlay and labeling ...............................................................................................................................2 - 11 Alarm Lamp ............................................................................................................................................2 - 12 Keys ......................................................................................................................................................2 - 13 Patient Cable Connectors .........................................................................................................................2 - 14 Audio Indicator ..............................................................................................................................................2 - 14 Pulse Tone Function..................................................................................................................................2 - 14 Multiple-Level Volume...............................................................................................................................2 - 14 Real-time Clock...............................................................................................................................................2 - 14 Standby Mode ...............................................................................................................................................2 - 15 Alarm Information ...........................................................................................................................................2 - 16 Basics ....................................................................................................................................................2 - 16 Alarm Notification ...................................................................................................................................2 - 16 Audio Alarm Pause, Audio Alarm Off, and Audio Alarm Silencing ................................................................2 - 16 Configuration management..............................................................................................................................2 - 16 Barcode scanner.............................................................................................................................................2 - 16 Introduction ....................................................................................................................................................3 - 2 Part Replacement ............................................................................................................................................3 - 2 Monitor Status Check ......................................................................................................................................3 - 2 Technical Alarm Check ....................................................................................................................................3 - 3 Troubleshooting Guide ....................................................................................................................................3 - 4 Power On/Off Failures ............................................................................................................................3 - 4

Accutorr V Service Manual

0070-10-0702

Table of Contents

Display Failure ........................................................................................................................................3 - 4 LED Digital Display and Indication Lamp Failure ..........................................................................................3 - 4 Alarm Problems ......................................................................................................................................3 - 5 Key Failure ............................................................................................................................................3 - 5 Recorder Failures.....................................................................................................................................3 - 6 Interface Failures ....................................................................................................................................3 - 6 Power Supply Failures ..............................................................................................................................3 - 6 Network Related Problems .......................................................................................................................3 - 7 Software Upgrade Problems ....................................................................................................................3 - 8 Bar Code Scanner Failures .......................................................................................................................3 - 8 NIBP Measurement Failures ......................................................................................................................3 - 9 SpO2 Measurement Failure ......................................................................................................................3 - 9 Temperature Measurement Failures ...........................................................................................................3 - 9 Tools .............................................................................................................................................................4 - 2 Preparation for Disassembly .............................................................................................................................4 - 2 Disassembly ..................................................................................................................................................4 - 3 Separating the Front and Rear Halves of the Monitor ..................................................................................4 - 3 Disassembling the Front Housing Assembly ................................................................................................4 - 6 Removing the Main Rack Assembly............................................................................................................4 - 9 Removing NIBP Module ...........................................................................................................................4 - 11 Removing the Main (CPU) Board ..............................................................................................................4 - 12 Removing the Power Board .......................................................................................................................4 - 13 Removing the Battery Connector Assembly..................................................................................................4 - 13 Removing the Fan and Speaker .................................................................................................................4 - 14 Removing the Power Socket .....................................................................................................................4 - 15 Disassembling the Temperature Module......................................................................................................4 - 16 Disassembling SpO2 Modules ...................................................................................................................4 - 17 Disassembling the NIBP Module ................................................................................................................4 - 20 Main Assembly...............................................................................................................................................5 - 2 Front Assembly (Color LCD)..............................................................................................................................5 - 3 Front Assembly (Black and White LCD) ..............................................................................................................5 - 8 Rear Housing Assembly ...................................................................................................................................5 - 12 Main Bracket Assembly ...................................................................................................................................5 - 13 NIBP Assembly (Pump M6Q-100003- - -) ...........................................................................................................5 - 16 NIBP Assembly (Pump 082-000056-00) ............................................................................................................5 - 19 Temperature Module Subassembly ....................................................................................................................5 - 22 Temperature Power Module Subassembly ..........................................................................................................5 - 24 Introduction ....................................................................................................................................................6 - 2 Test Report..............................................................................................................................................6 - 3 Recommended Frequency .........................................................................................................................6 - 3 Visual Test......................................................................................................................................................6 - 4 Power-on Test .................................................................................................................................................6 - 4 NIBP Calibration.............................................................................................................................................6 - 4 NIBP Accuracy Test .........................................................................................................................................6 - 8 NIBP Leakage Test ...................................................................................................................................6 - 9 SpO2 Test ......................................................................................................................................................6 - 12 SpO2 Test Under Normal Conditions .........................................................................................................6 - 12 SpO2 Test in Motion Mode .......................................................................................................................6 - 12 Summary of Test Methods .........................................................................................................................6 - 13 Testing the Optional Temperature Module..........................................................................................................6 - 14 Nurse Call Performance Test ............................................................................................................................6 - 16 Bar Code Scanner Test ....................................................................................................................................6 - 18

ii

0070-10-0702

Accutorr V Service Manual

Table of Contents

Electrical Safety Tests ......................................................................................................................................6 - 20 Enclosure Leakage Current Test .................................................................................................................6 - 20 Earth Leakage Current Test .......................................................................................................................6 - 20 Patient Leakage Current Test .....................................................................................................................6 - 21 Patient Auxiliary Leakage Current Test........................................................................................................6 - 21 Recorder Check ..............................................................................................................................................6 - 22 Software upgrade ...........................................................................................................................................6 - 23

Accutorr V Service Manual

0070-10-0702

iii

Table of Contents

This page intentionally left blank.

iv

0070-10-0702

Accutorr V Service Manual

Preface

Introduction

Preface
This manual provides detailed information about the assembling, disassembling, testing, and troubleshooting of the equipment to support effective troubleshooting and repair. It is not intended to be a comprehensive, in-depth explanation of the product architecture or technical implementation. Observance of the manual is a prerequisite for proper equipment maintenance and prevents equipment damage and personnel injury. Refer to the Accutorr V Operating Instructions: P/N 0070-00-0699-XX for information for operating this instrument. This manual is based on the maximum configuration. Therefore, some contents may not be applicable. For questions, contact Service. This manual is for biomedical engineers, authorized technicians, or service representatives responsible for troubleshooting, repairing, and maintaining the monitors. Contents of this manual are subject to change without prior notice. A password is required to access the service mode. The service password is 321. The Accutorr V configurations are: Accutorr V with Nellcor Pulse Oximetry includes NIBP, Nellcor SpO2, a Liquid Crystal Display (LCD), and Recorder Accutorr V with Nellcor Pulse Oximetry and SmarTemp includes NIBP, Nellcor SpO2, SmarTemp, a Liquid Crystal Display (LCD), and Recorder Accutorr V with Masimo SET Pulse Oximetry includes NIBP, Masimo SpO2, a Liquid Crystal Display (LCD), and Recorder Accutorr V with Masimo SET Pulse Oximetry and SmarTemp includes NIBP, Masimo SpO2, SmarTemp, a Liquid Crystal Display (LCD), and Recorder Accutorr V with DPM Pulse Oximetry includes NIBP, DPM SpO2, a Liquid Crystal Display (LCD), and Recorder Accutorr V with DPM Pulse Oximetry and SmarTemp includes NIBP, DPM SpO2, SmarTemp, a Liquid Crystal Display (LCD), and Recorder Accutorr V with NIBP and SmarTemp includes NIBP, SmarTemp, a Liquid Crystal Display (LCD), and Recorder Accutorr V with NIBP only includes NIBP, a Liquid Crystal Display (LCD), and Recorder All Accutorr V configurations can be upgraded with a barcode scanner. Masimo Patents: This device (MASIMO SpO2 Module) is covered under one or more of the following U.S. Patents 5,758,644, 5,823,950, 6,011,986, 6,157,850, 6,263,222, 6,501,975, and other applicable patents listed at: www.masimo.com/patents.htm. Possession or purchase of this device does not convey any express or implied license to use the device with replacement parts which would, alone, or in combination with this device, fall within the scope of one or more of the patents relating to this device.

Accutorr V Service Manual

0070-10-0702

Introduction

Preface

Nellcor Patents: This device (Nellcor SpO2 Module) is covered under one or more of the following U.S. Patents Patent No. 5,485,847, 5,676,141, 5,743,263, 6,035,223, 6,226,539, 6,411,833, 6,463,310, 6,591,123, 6,708,049, 7,016,715, 7,039,538, 7,120,479, 7,120,480, 7,142,142, 7,162,288, 7,190,985, 7,194,293, 7,209,774, 7,400,919, and 7,212,847. Possession or purchase of this device does not convey any express or implied license to use the device with replacement parts which would, alone, or in combination with this device, fall within the scope of one or more of the patents relating to this device.

vi

0070-10-0702

Accutorr V Service Manual

Warnings, Cautions, and Notes

Introduction

Warnings, Cautions, and Notes


Read and adhere to all of the warnings and cautions listed throughout this manual. A WARNING is provided to alert the user to potentially serious outcomes (death, injury or serious adverse events) to the patient or the user. A CAUTION is provided to alert the user that special care should be taken for the safe and effective use of the device. They will include actions to be taken to avoid effects on patients or users that will not be potentially life threatening or result in serious injury, but about which the user should be aware. A NOTE is provided when additional general information is available.

Warnings
WARNING: Internal Electrical Shock Hazard - This unit does not contain any user-serviceable parts. Do not remove instrument covers. Refer servicing to qualified personnel. When the integrity of the protective earth conductor, in the installation or its arrangement, is in doubt, the equipment should be operated from its internal battery. Observe all CAUTION and WARNING labels on the unit. WARNING: Possible explosion hazard. Do not operate machine near flammable anesthetic agents or other flammable substances. Do not use flammable anesthetic agents (i.e., ether or cyclopropane.) WARNING: Always place the unit on a flat, rigid surface or onto a Mindray approved stable mounting bracket. WARNING: To ensure proper performance and safety and to prevent the voiding of the warranty, only use authorized parts and accessories with the Accutorr V. Use of unauthorized accessories may result in erroneous readings. WARNING: Use only cuffs with approved quick connect type connectors. WARNING: The Accutorr V is not intended for use in a magnetic resonance imaging (MRI) environment and may interfere with MRI procedures. WARNING: Danger of explosion if battery is incorrectly replaced. Replace only with the same or equivalent type recommended by the manufacturer. Dispose of used batteries according to the manufacturers instructions and local regulations. Batteries used in this device may present a risk of fire or chemical burn if mistreated. Do not incinerate battery, possible explosion may occur. WARNING: Do not use a damaged or broken unit or accessory. WARNING: Operation of the Accutorr V below the minimum amplitude or value of patient physiological signal may cause inaccurate results.

Accutorr V Service Manual

0070-10-0702

vii

Introduction

Cautions

WARNING: Use of accessories, transducers, and cables other than those specified in the manual may result in increased Electromagnetic Emissions or decreased Electromagnetic Immunity of the Accutorr V. It can also cause delayed recovery after the discharge of a cardiac defibrillator. WARNING: Perform the decontamination or cleaning process with the unit powered down and power cord removed. WARNING: Electrical safety tests are a proven means of verifying the electrical safety of the monitor. They are intended for determining potential electrical hazards. Failure to identify these hazards in a timely manner may cause personnel injury. WARNING: Commercially available test equipment such as a safety analyzer can be used for electrical safety tests. Verify that the test equipment can be safely and reliably used with the monitor before use. The service personnel should acquaint themselves with the use of the test equipment. WARNING: Electrical safety tests should meet the requirements of the latest editions of EN 60601-1 and UL 60601. WARNING: These electrical safety tests do not supersede local requirements. WARNING: All devices using the AC mains and connected to medical equipment within patient environments must meet the requirements of the IEC 60601-1-1 medical electrical systems standard and should be put under electrical safety tests at the frequency recommended for the monitor.

Cautions
CAUTION: Observe extreme caution when a defibrillator is in use. Do not touch any part of the patient, table, or monitor when a defibrillator is in use. The Accutorr V should not be used adjacent to or stacked with other equipment. If adjacent or stacked use is necessary, the Accutorr V should be observed to verify normal operation in the configuration in which it will be used. The unit should be checked periodically for obstructed vents. If an obstruction is found, refer the unit to qualified service personnel. At the end of their life, dispose of the Accutorr V, accessories, and single use supplies in accordance with local regulations. Dispose of packaging waste in accordance with local regulations. When equipped with Nellcor SpO2, use only Nellcor oxygen transducers including Nellcor Oxisensor patient dedicated adhesive sensors. Use of other oxygen transducers may cause improper oximeter performance. When equipped with MASIMO SpO2, use only MASIMO oxygen transducers including MASIMO LNOP and MASIMO LNCS patient dedicated adhesive sensors and MASIMO PC Series Patient Cable. Use of other oxygen transducers may cause improper Oximetry performance.

CAUTION:

CAUTION:

CAUTION:

CAUTION:

viii

0070-10-0702

Accutorr V Service Manual

Cautions

Introduction

CAUTION:

Inaccurate readings may be caused by incorrect sensor application or use; significant levels of dysfunctional hemoglobins (i.e. carbohemoglobins or methemoglobin); or intra-vascular dyes such as indocyanine green or methylene blue; exposure to excessive illumination, such as surgical lamps (especially ones with a Xenon light source), bilirubin lamps, fluorescent lights, infrared heating lamps, or direct sunlight; excessive patient movement; venous pulsations; electro-surgical interference; and placement of a sensor on an extremity that has a blood pressure cuff, arterial catheter, or intra-vascular line. Route cables neatly. Ensure cables, hoses, and wires are kept away from patients neck to avoid strangulation. Keep floors and walkways free of cables to reduce risk to hospital personnel, patients, and visitors. If the sensor or patient cable is damaged in any way, discontinue use immediately. When cleaning sensors, do not use excessive amounts of liquid. Wipe the sensor surface with a soft cloth, dampened with the cleaning solution. To prevent damage, do not soak or immerse the sensor in any liquid solution. DO NOT ATTEMPT TO STERILIZE. Recharge the Lithium ion battery while in the unit at room temperature. If using the Accutorr V in a hot environment, the Lithium ion battery may not charge when the unit is connected to the AC mains. Remove the battery if the Accutorr V is not used for an extended period of time. The Communications Connectors on the Accutorr V are only for use with IEC 60601-1-1 compliant equipment. Never place fluids on top of this monitor. If fluid spills on the unit, wipe clean immediately and refer the unit to qualified service personnel. Before disassembling the monitor, eliminate static charges. When disassembling the parts labeled with static-sensitive symbols, wear electrostatic discharge protection such as antistatic wristband or gloves. Follow the correct sequence to disassemble the monitor. Otherwise, the monitor may be damaged permanently. Disconnect all the cables before disassembling any parts. Take care not to damage any cables or connectors. Properly connect the cables or wires when reassembling the monitor to avoid short circuit. When assembling the monitor, select proper screws. If the wrong size screw is tightened by force, the monitor may be damaged and the screw or the part may not function as expected. All tests should be performed by qualified personnel only. Disconnect the monitor from the patient and make sure that important data is saved before upgrading the monitor. Do not shut down or power off the equipment when upgrading the bootstrap program. Otherwise, it may cause the equipment to break down.

CAUTION:

CAUTION:

CAUTION:

CAUTION:

CAUTION:

CAUTION:

CAUTION:

CAUTION:

CAUTION: CAUTION:

CAUTION:

Accutorr V Service Manual

0070-10-0702

ix

Introduction

Notes

CAUTION:

Program upgrades should be performed by qualified service personnel only.

Notes
NOTE: The Accutorr V should be operated only by trained and qualified personnel. Use disposable and single use accessories only once. Place the equipment in a location where the screen can easily be seen and the operating controls can easily be accessed. The instructions in this manual are based on the maximum configuration. The optional Temperature module kit must be installed only by trained personnel, and proper ESD prevention methods must be followed. Only devices specified by Mindray DS USA, Inc. shall be connected the RS-232 port When the RS-232 connector is used for DIAP, barcode power must be set to OFF. Disconnect the Accutorr V from the mains to isolate it from the mains power during an emergency. NOTE: NOTE:

NOTE:

NOTE:

NOTE:

NOTE:

NOTE:

Safety Designations
Safety designations per IEC 60601-1 Standard: Type of protection against electric shock Class 1 with internal electric power source. Where the integrity of the external protective earth (ground) in the installation or its conductors is in doubt, the equipment shall be operated from its internal electric power source. Degree of protection against electric shock Monitor - Type B applied part. NIBP - Type BF defibrillation protected applied part. SpO2 - Type BF protected applied part. Temp - Type BF protected applied part. Supply Connection 100 240 VAC 50/60 Hz 0.85 0.5 A Mode of Operation Continuous

0070-10-0702

Accutorr V Service Manual

Product Limitations

Introduction

Protection Against Hazard of Explosion Protection Against Ingress of Liquids Degree of Electrical Connection Between Equipment and Patient Degree of Mobility

Not Protected (Ordinary) IPX1 Equipment designed for direct electrical and non-electrical connection to the patient. Portable

Product Limitations
Non-invasive blood pressure (NIBP) accuracy depends on the application of the proper cuff size. See Chapter 3.0 of the Operating Instructions for detailed information. The Accutorr V will not operate effectively on patients who are experiencing convulsions or tremors. The Accutorr V is a portable device intended for intra-hospital use. If the pressure cuff is not placed at the patients heart level, the NIBP measurement may be subject to error, due to the hydrostatic effect. The pulse rate data displayed on the Accutorr V is computed from the measurement of peripheral pulses (peripheral pulses taken only during a measurement cycle). The rate measured by the Accutorr V may differ from the rate of an ECG monitor. This is because the ECG is an electrical signal that may not always result in a peripheral pulse. Administration of certain vasoconstrictor drugs (for example, norepinephrine), may reduce peripheral perfusion to a level that prevents the Accutorr V from taking pulse rate measurements. Arterial compression, tricuspid regurgitation, or other conditions may reduce perfusion to a level that prevents the Accutorr V from taking pulse rate measurements. The presence of arrhythmias may increase the time required to complete a measurement and may extend this time so that a measurement cannot complete. The Accutorr V is not intended for use during CPR. The monitor uses an oscillometric technique based on normal peripheral circulation to compute blood pressure.

Unpacking
Remove the instrument from the shipping carton and examine it for signs of shipping damage. Save all packing materials, invoice, and bill of lading. These may be required to process a claim with the carrier. Check all materials against the packing list. Contact the Customer Service Department (800) 288-2121 or (201) 265-8800 for prompt assistance in resolving shipping problems.
NOTE: The Accutorr V should only be shipped in its original packing materials to avoid shipping damage.

Accutorr V Service Manual

0070-10-0702

xi

Introduction

Symbols and Descriptions

Symbols and Descriptions


SYMBOL DESCRIPTION SYMBOL DESCRIPTION

Attention, Consult Accompanying Documents / Refer to Manual Equipotentiality Equipotential grounding Defibrillator-proof Type BF Equipment Type BF Equipment

Alternating Current (AC)

Adult

T1 SpO2

Predictive Thermometer Connector SpO2 Connector Electrical connectors

Pediatric/Child

Neonate

Operating on battery power

Battery Connected to AC mains

Manufacturer

NIBP Connector

Power On/Off Standby

Recycle

Serial number

Up key

REF

Part Number

Confirm key

Patient Information key

Down key

Main menu key

Deflate Cuff key

xii

0070-10-0702

Accutorr V Service Manual

Symbols and Descriptions

Introduction

SYMBOL

DESCRIPTION

SYMBOL

DESCRIPTION

Set alarms key

Patient Size key

Start NIBP key

Alarm Silence key

Display Tabular Trends/Pleth Wave NIBP interval key

NC1

Nurse Call connector RS-232 connector (Serial Port 1) Network connector Alarm Disabled indicator on LCD display Audio Alarm Off indicator on LCD display

SP1

Print key (front panel)

CS1

Print key (recorder) Alarm Silenced indicator on LCD display

Classified by Underwriters Laboratories Inc. with respect to electric shock, fire and mechanical hazards, only in accordance with UL 60601-1, CAN/CSA C22.2 NO.601-1, IEC 60601-1-1, IEC 60601-2-30, IEC 60601-2-49.

Accutorr V Service Manual

0070-10-0702

xiii

Introduction

Symbols and Descriptions

This page intentionally left blank.

xiv

0070-10-0702

Accutorr V Service Manual

1.0

Theory of Operation

Introduction.................................................................................................... 1-2 Controls and Indicators ................................................................................... 1-2 System Overview............................................................................................ 1-2 Hardware Structure......................................................................................... 1-4

Accutorr V Service Manual

0070-10-0702

1-1

Introduction

Theory of Operation

1.1

Introduction
The Accutorr V monitors the following patient vital signs: non-invasive blood pressure (NIBP), pulse oxygen saturation (SpO2), pulse rate (PR), and temperature (Temp) for a single adult, pediatric, or neonatal patient. Temperature is measured using the optional Temperature Module.

1.2 1.3

Controls and Indicators


For information on controls, connectors, and indicators, refer to the Accutorr V Operating Instructions, Part Number 0070-10-0699-02.

System Overview
FIGURE 1-1 shows the relationship of the Accutorr V monitoring systems mechanical, hardware, and software components. FIGURE 1-2 shows the Accutorr V monitoring systems PC board connections. SYSTEM

MECHANICAL

HARDWARE

SOFTWARE SYSTEM SOFTWARE

MAIN UNIT

MAIN BOARD

RECORDER

POWER BOARD

PARAMETER SOFTWARE

BATTERY CHAMBER TEMP MODULE CHAMBER BARCODE SCANNER SUPPORT

KEY & DISPLAY

UPGRADE SOFTWARE

PARAMETER BOARDS

PRINT SOFTWARE

RECORDER BOARD

GATEWAY SOFTWARE

BARCODE SCANNER

FIGURE 1-1 System Mechanical, Hardware, and Software Overview

1-2

0070-10-0702

Accutorr V Service Manual

Theory of Operation

System Overview

PARAMETER RECEPTACLE

J1 SpO2 J2

J2 TEMP J4 NIBP J2

P2 RECORDER J2 ISOLATION POWER BOARD P1 J6

J4 ISOLATION POWER BOARD J3

J2 J1 J1 J7 POWER P3 P5 J6

J2

J3 MAIN BOARD J8 J9 J10

J4

J5

J11

J12

J13

P4 NETWORK CONNECTOR BDM PORT SPEAKER

RS232

J3

J4

J5

KEY & DISPLAY BOARD P2

NURSE CALL

AC INLET

FAN

BATTERY LCD

FIGURE 1-2 The System PCBA Connections

Accutorr V Service Manual

0070-10-0702

1-3

Hardware Structure

Theory of Operation

1.4

Hardware Structure
FIGURE 1-3 shows the Accutorr Vs Hardware Structure. The core of the system is the main board, which supplies power for all parameter modules. The parameter modules directly communicate with the main board. The the main board processes measurements and the status of all modules, and displays them through the key and display board.

NIBP MODULE

SpO2 MODULE

TEMP MODULE

ISOLATION POWER BOARD COMMUNICATIONS /POWER SUPPLY LED & KEY KEY & DISPLAY BOARD

ISOLATION POWER BOARD COMMUNICATIONS /POWER SUPPLY

ETHERNET

LCD

MAIN BOARD

RS232

POWER SUPPLY

NURSE CALL

SPEAKER

RECORDER

BATTERY FAN POWER BOARD AC

ACCUTORR V FIGURE 1-3 The Accutorr Vs Hardware Structure


NOTE: The SpO2 isolation power board does not apply to models with DPM Pulse Oximetry.

1-4

0070-10-0702

Accutorr V Service Manual

Theory of Operation

Hardware Structure

1.4.1

Main Board
FIGURE 1-4 shows the main board, which provides resources and support for the overall system. It controls the LCD, LED, keyboard, speaker, and recorder. It also communicates with parameter modules and connectors. The main board also controls communication with the speaker, the recorder, and all external connectors.

WATCHDOG RTC E2 PROM

ETHERNET NURSE CALL SPEAKER

PHY RTL8201

CPU
AUDIO PROCESS CIRCUIT

SERIAL PORT 0 RS232 IC

RS232 SERIAL PORT 1: NIBP

SDRAM 1.5V

FLASH

SERIAL PORT 2: SpO2

FPGA

SERIAL PORT 3: TEMP SERIAL PORT 4: RECORDER

LINEAR POWER

3.3V

5.0V

12V

LCD

LED

KEY

FIGURE 1-4 Main Board Principle Diagram The main board communicates with all parameter modules and the recorder through an FPGA extended serial port. The main board supplies information using FPGA to the key and display board. It drives the display, detects the keys, and implements the user interface. The main board controls the alarm indicator using FPGA. The main board controls the speaker to give audible alarms, key tones, and Pitch Tone. The main board provides the nurse call connection, network connection, and R232 connection. The real-time clock is implemented by the RTC chip. The RT clock is powered by the AC mains, battery, or button cell on the main board. The button cell ensures the continuous working of the clock in the event that the AC mains and batteries are not available. SDRAM stores running program instructions and data temporarily. The system memory and trend data memory is flashed. The device configuration memory is EEPROM.

Accutorr V Service Manual

0070-10-0702

1-5

Hardware Structure

Theory of Operation

1.4.2

Power Board
The power board, shown in FIGURE 1-5, converts the input power (AC mains or battery) to different working voltages for other boards. It also charges the battery.
16.8V AC INPUT EMI FILTER RECTIFIER & FILTER FLYBACK CONVERTER RECTIFIER & FILTER DC/DC CONVERTER 12V OUTPUT

0VP&0CP

DC/DC CONVERTER CURRENT DETECTION PMW CONTROLLER COUPLER FEEDBACK & ISOLATION CIRCUIT CHARGING CIRCUIT 0VP&0PP

5V OUTPUT

BATTERY

DC/DC CONVERTER

3.3V OUTPUT

SWITCH CIRCUIT

0VP&0CP

PCON

FIGURE 1-5 Power Board Principle Diagram The AC flows through the EMI filter and the rectifier and filter. The rectifier and filter converts the AC to 16.8V DC voltage by the Flyback converter. The 16.8V DC voltage is the main input to the DC/DC converters and charging circuit. The DC/DC converters convert 16.8V DC to 12V, 5V, and 3.3V DC. The charging circuit charges the lithium battery. When the Accutorr V is not connected to the AC mains, the battery supplies power to the DC/DC converters. The 16.8V DC output is protected against over-voltage and over-power. The 12V, 5V, and 3.3V DC outputs are protected against over-voltage, short-circuit, and overcurrent.

1-6

0070-10-0702

Accutorr V Service Manual

Theory of Operation

Hardware Structure

1.4.3

Key and Displays Board


The key and displays board, shown in FIGURE 1-6, provides the user interface. The board contains the LCD module, 7-segment digital display, LED indication lamp, and keys.

LCD SIGNAL

FSTN LCD MODULE

LED

CPLD

LED&KEY

3.3V DISABLE

START-UP CIRCUIT

ADV 3.3V VSB

KEY MATRIX

Switch

DETECTION

FIGURE 1-6 Key and displays Board Principle Diagram The LCD module adjusts contrast and brightness (only for black-and-white LCD display).
NOTE: The monitor with color LCD does not have a Contrast setting item in the common setup menu.

The 7-segment digit units display parameter data. The AC indicator is driven by the ADV output from the power board, and the working status indicator (built in the Power On/Off key) is driven by 3.3V voltage. The battery indicator is jointly controlled by the flash control signal, ADV signal, and VBC signal. The keypad contains the power ON/OFF key and the other 13 function keys.

Accutorr V Service Manual

0070-10-0702

1-7

Hardware Structure

Theory of Operation

1.4.4 1.4.4.1

Parameter Boards SpO2 Module


FIGURE 1-7 shows the SpO2 module parameter board diagram.
/RST WATCHDOG WDI 4KV ISOLATED POWER SUPPLY ISOLATED SERIAL IN/OUTPUT PORT HOST DRAM DRIVE DAC +12V

CPU
FLASH LED DRIVE CIRCUITRY

+3.3V +5V -5V

SENSOR A/D CONVERTER GAIN CONTROL CIRCUITRY

OFFSET DAC

PROGRAMMABLE GAIN CIRCUITRY

OFFSET AMPLIFIER

FIGURE 1-7 SpO2 Module Principle Diagram The SpO2 sensor collects the pulsing red and infrared light signals transmitting through the finger or toe, and processes the collected signals to create the measured result. The SpO2 module controls the LED drive circuit and the amplifying circuit gain corresponding to the finger or toe perfusions and transmittances.

1-8

0070-10-0702

Accutorr V Service Manual

Theory of Operation

Hardware Structure

1.4.4.2

NIBP Module
FIGURE 1-8 shows the NIBP module parameter board diagram.
PRESSURE SIGNAL AMPLIFIER CIRCUITS FOR PROTECT

CUFF

PRESSURE SENSOR FOR PROTECT

OVER PRESSURE PROTECT CIRCUIT

WATCHDOG

CPU

ASYCHRONOUS SERIAL COMMUNICATION

PRESSURE SENSOR

PRESSURE SIGNAL AMPLIFIER CIRCUITS

PRESSURE SIGNAL

A/D CONVERTER

MANIFOLD

MOTOR CONTROL FEEDBACK SIGNAL

VALVES AND PUMP

PUMP AND VALVE CONTROL CIRCUIT

FLASH

DRAM

FIGURE 1-8 NIBP Module Principle Diagram The Accutorr V calculates NIBP values using the oscillometric method of noninvasive blood pressure measurement. These measurements correspond to comparisons with auscultatory values, measured using the fifth Korotkoff sound within ANSI/AAMI SP10 standards for accuracy.

1.4.4.3

Optional Temperature Module


FIGURE 1-9 shows the Temperature module parameter board diagram.

PROBE RECOGNIZE CIRCUIT

TEMPERATURE DETECT CIRCUIT

A/D CIRCUIT

CPU SYSTEM (ROM & RAM WATCHDOG)

POWER & SIGNAL ISOLATE CIRCUIT

MAIN BOARD

PROBE HEAT CIRCUIT

FIGURE 1-9 Optional Temperature Module Principle Diagram The Temperature Module uses a thermistor as a sensor for measuring temperature. The resistance of a given thermistor is nonlinearly relative to the temperature. The Temperature Module measures the resistance of the thermistor and converts it into temperature.

Accutorr V Service Manual

0070-10-0702

1-9

Hardware Structure

Theory of Operation

1.4.5

Recorder
The recorder receives data from the main board and sends it to the thermal printhead for printing. It has a button to start or stop printing and a green LED to indicate the presence or absence of paper.

1 - 10

0070-10-0702

Accutorr V Service Manual

Theory of Operation

Hardware Structure

1.4.6

Bar Code Scanner


The bar code scanner reads one-dimensional and two-dimensional bar codes to simplify admitting a patient. The bar code scanner communicates with the monitor. The serial port supplies it with power. The scanners serial port is defined in the following table:
PIN 2 3 5 9 DEFINITION Barcode_RX Barcode_TX GND VCC5VDC

The monitors serial port is defined in the following table:


PIN 2 3 5 9 DEFINITION Monitor_TX Monitor_RX GND VCC5VDC

Basic settings of the bar code scanner are listed in the following table:
HOST PARAMETERS Baud Rate Data Bits Stop Bits Calibration bit Handshaking BAR CODE SCANNER 9600 8 1 0 None FACTORY DEFAULT 9600 8 1 0 None

USER PARAMETERS Beeper Tone Beeper Volume Trigger Mode Parameter Scanning

BAR CODE SCANNER Medium Medium Level Disable

FACTORY DEFAULT Medium High Auto Aim Enable

DATA FORMAT Prefix Value Suffix 1 Value Suffix 2 Value Scan Data Transmission Format

BAR CODE SCANNER 7013 <CR><LF> 7013 <CR><LF> <PREFIX><DATA><SUFFIX 1><SUFFIX 2>

FACTORY DEFAULT 7013 <CR><LF> 7013 <CR><LF> Data only

Accutorr V Service Manual

0070-10-0702

1 - 11

Hardware Structure

Theory of Operation

To change parameter settings, scan the following bar codes in sequence. 1. Set parameter defaults:

2. Set beeper volume:

3. Set trigger mode:

4. Set scan data transmission format:

1 - 12

0070-10-0702

Accutorr V Service Manual

Theory of Operation

Hardware Structure

5. Disable parameter scanning:

Accutorr V Service Manual

0070-10-0702

1 - 13

Hardware Structure

Theory of Operation

This page intentionally left blank.

1 - 14

0070-10-0702

Accutorr V Service Manual

2.0

Specifications

Introduction.................................................................................2-2 NIBP Measurement ......................................................................2-2 NIBP Measurement Cycle .............................................................2-4 SpO2 Measurement.....................................................................2-6 Temperature ................................................................................2-10 Display Area, Indicator, and Controller ..........................................2-10 Audio Indicator ...........................................................................2-14 Real-time Clock ...........................................................................2-14 Standby Mode ............................................................................2-15 Alarm Information........................................................................2-16 Configuration management ..........................................................2-16 Barcode scanner .........................................................................2-16

Accutorr V Service Manual

0070-10-0702

2-1

Introduction

Specifications

2.1

Introduction
The Accutorr V monitors patient vital signs (including non-invasive blood pressure (NIBP), pulse oxygen saturation (SpO2), and pulse rate (PR)) for a single adult, pediatric, or neonatal patient. It also monitors oral or rectal mode temperature (Temp) for a single adult or pediatric patient; or axillary mode temperature for a single neonate, adult, or pediatric patient using the optional Temperature Module.

2.2

NIBP Measurement
The NIBP module is used to measure the systolic pressure, diastolic pressure, and mean pressure in the neonatal, pediatric, and adult modes. The Accutorr V calculates NIBP values using the oscillometric method of noninvasive blood pressure measurement. These measurements correspond to comparisons with auscultatory values, measured using the fifth Korotkoff sound within ANSI/AAMI SP10 standards for accuracy.

2.2.1

Blood Pressure
PARAMETER Systolic pressure measurement PATIENT SIZE Adult Pediatric Neonate Adult Pediatric Neonate Adult Pediatric Neonate Mean error: <5 mmHg Standard deviation: <8 mmHg RANGE 55 235 mmHg 55 160 mmHg 45 120 mmHg 30 200 mmHg 30 150 mmHg 20 100 mmHg 30 235 mmHg 30 160 mmHg 20 120 mmHg

Diastolic pressure measurement

Mean pressure measurement

Measurement accuracy

Measurements outside of the stated ranges are not guaranteed to be accurate.

2.2.2

Static Accuracy
Measurement range: Static pressure measurement accuracy: 0 300 mmHg 3 mmHg

2.2.3

Maximum Cuff Pressure


Normal Use Over Pressure Protection In normal use, the over-pressure detection is controlled by software. Once the cuff pressure exceeds the threshold, the software enables the system to deflate the cuff. Adult Pediatric Neonate 300 mmHg 200 mmHg 150 mmHg

2-2

0070-10-0702

Accutorr V Service Manual

Specifications

NIBP Measurement

Single Fault Over-pressure protection In single fault conditions, the hardware controls the cuff deflation to prevent the cuff pressure from exceeding the following ranges: Adult Pediatric Neonate 330 mmHg 220 mmHg 165 mmHg

2.2.4

Cuff Inflation
The Non-Invasive Blood Pressure inflation source brings a volume of 200 ccs to a pressure of 300 mmHg in less than or equal to 10 seconds

2.2.5

Maximum Leakage
The Non-Invasive Blood Pressure Cuff driver allows a pressure drop to be, at most, 1 mmHg in 10 seconds as measured with a 200 cc volume at differential pressures of 250 mmHg, 150 mmHg, and 50 mmHg.

2.2.6

Deflation Rate
The Non-Invasive Blood Pressure Cuff venting mechanism in adult mode vents a volume of 500 cc, reduced from a pressure of 260 mmHg to a pressure of 15 mmHg in a maximum of 10 seconds. The Non-Invasive Blood Pressure Cuff venting mechanism in neonatal mode vents a volume of 500 cc, reduced from a pressure of 150 mmHg to a pressure of 5 mmHg in a maximum of 5 seconds.

2.2.7 2.2.8

Update Period
The Non-Invasive Blood Pressure Parameter has an update period of one (1) second.

Initial NIBP Pressure Setup and Range


The initial pressure is adjustable, and should be set to the following default values:

PATIENT TYPE Adult Pediatric Neonate

DEFAULT INITIAL PRESSURE 180 mmHg 140 mmHg 100 mmHg

ADJUST RANGE 100 280 mmHg 60 180 mmHg 40 120 mmHg

Manual pressure changes in increments of 5 mmHg.

Accutorr V Service Manual

0070-10-0702

2-3

NIBP Measurement Cycle

Specifications

2.3

NIBP Measurement Cycle


There are two different modes of measurement operation: manual and interval modes. The manual mode requires the operator to initiate the measurement cycle. The interval mode follows a configured plan of automatically initiated measurement cycles, using the follow selections: OFF (manual), STAT, or 1, 2, 3, 5, 10, 15, 20, 30, 60, 120, 240 minutes
NOTE: When the NIBP STAT interval is selected, the Accutorr V takes back to back (one right after the other) blood pressure readings. As a safety precaution, there is a five minute or 10 measurement limit for continuous NIBP measurements. After 5 minutes or 10 measurements, the NIBP module automatically switches to the mode in use before NIBP STAT was selected. This reduces the chance of surface vessel rupture (petechia).

The maximum adult and pediatric measurement cycle is 180 seconds. The maximum neonatal measurement cycle is 90 seconds. The NIBP module adjusts the inflation pressure automatically according to the systolic pressure of the last measurement. If the first measurement is unsuccessful, the subsequent inflation pressures change to 50 mmHg higher than the previous systolic pressure in adult or pediatric mode, and 40 mmHg higher than the previous systolic pressure in neonate mode. (The subsequent inflation pressure should be greater than the systolic pressure in the previous measurement).

2.3.1

Cuff Inflation Time


If the cuff pressure does not attain 15 mmHg within 20 seconds for adult and pediatric patients, or 15 mmHg within 10 seconds for neonate patients from the start of inflation, the Accutorr V retries measuring up to three times. If the target pressure is not reached after three retries, the cuff is deflated, and a status code displays.

2.3.1.1

Cuff Pressure Automatic Check Algorithm


If the monitor cannot inflate the cuff to the Pressure Threshold within the expected time period, the monitor deflates the cuff, cancels the measurement, shuts off the pump, and displays the message Unable to Measure. The cancellation occurs based on the following tables.

OPERATING MODE Adult and Pediatric Neonate

PRESSURE THRESHOLD 15 mmHg 15 mmHg

DURATION 20 sec. 10 sec.

NUMBER OF RETRIES 3 3

OPERATING MODE Adult, Pediatric, Neonate

TARGET INFLATION PRESSURE 60 mmHg

NUMBER OF RETRIES 3

2-4

0070-10-0702

Accutorr V Service Manual

Specifications

NIBP Measurement Cycle

2.3.2

Automatically Adjusted Inflation Value after First Measurement


In the Interval and Stat modes, the inflation value of the next measurement will be automatically adjusted by the NIBP module based on the systolic value of the previous reading. After the first unsuccessful measurement, the subsequent inflation value will be the previous systolic pressure reading plus P2: Adult Mode: Pediatric Mode: Neonate Mode: P2 = 50 10 mmHg P2 = 50 10 mmHg P2 = 40 10 mmHg

2.3.3

Pulse Rate
Accuracy: 3 BPM or 3%, whichever is greater. Resolution: 1 BPM
PATIENT TYPE Adult Pediatric Neonate PULSE RANGE 35 245 BPM 35 245 BPM 70 245 BPM

Measurements outside of the stated ranges are not guaranteed to be accurate.

Accutorr V Service Manual

0070-10-0702

2-5

SpO2 Measurement

Specifications

2.4

SpO2 Measurement
When configured with the DPM SpO2 module, NELLCOR NELL-3 SpO2 module, or MASIMO MS-2013 SpO2 module, the Accutorr V can perform SpO2 measurements. The SpO2 measurement function complies with the requirements of ISO9919.

2.4.1

DPM SpO2 Module Performance


The DPM SpO2 accuracy has been validated in human studies against arterial blood sample reference measured with a CO-oximeter. Pulse oximeter measurements are statistically distributed, and only about two-thirds of the measurements can be expected to fall within the specified accuracy compared to CO-oximeter measurements.

2.4.1.1

SpO2
Range: 0 100% Resolution: 1% Accuracy in the static condition:
SPO2 RANGE PATIENT TYPE Adult or pediatric Neonate 70% TO 100% 2 digits 3 digits 0 69% Undefined Undefined

2.4.1.2

Pulse Rate
Range: 20 254 BPM Resolution: 1 bpm Accuracy in the static condition:
PATIENT TYPE Adult / Pediatric / Neonate PR RANGE 20 254 bpm ACCURACY 3 digits

2.4.1.3

Alarm setting range


SPO2 MODULE DPM SpO2 module SPO2 50 100% PR 0 254 bpm

2-6

0070-10-0702

Accutorr V Service Manual

Specifications

SpO2 Measurement

2.4.2

Masimo MS-2013 SpO2 Performance


This section details the performance measurements of the Masimo MS-2013, which is designed in accordance with the requirements of the interface of the Masimo MS-2013 board. The Accutorr V interface complies with the requirements of the Masimo MS-2013 board communication protocol.

2.4.2.1

SpO2
Compatible Sensors: See Section 6.6 for a list of compatible sensors. Measurement range: 1 100% Resolution: 1% Accuracy no motion conditions:
PATIENT TYPE Adult Pediatric Neonate SPO2 RANGE: 70% 100% 2 digits 2 digits 3 digits SPO2 RANGE: 1% 69% Undefined Undefined Undefined

Accuracy During motion conditions:


PATIENT TYPE Adult Pediatric Neonate SPO2 RANGE: 70% 100% 3% 3% 3% SPO2 RANGE: 1% 69% Undefined Undefined Undefined

SpO2 response time: Under the condition that the PR is 75 BPM and the averaging is 8 seconds, the maximum response time for the SpO2 value to increase from 60% to 95% is 20 seconds. Low Perfusion Performance:
LOW PERFUSION CONDITION PULSE AMPLITUDE > 0.02% % LIGHT TRANSMISSIBILITY > 5% SPO2 ACCURACY 2 digits PR ACCURACY 3 digits

Accutorr V Service Manual

0070-10-0702

2-7

SpO2 Measurement

Specifications

2.4.2.2

Pulse Rate
Resolution: 1 BPM Update frequency: 1 Hz Range and Accuracy:

PATIENT TYPE Adult/Pediatric/Neonate 1

PR RANGE 25 240 BPM

ACCURACY IN STATIC CONDITION1 3 digits

ACCURACY IN MOTION CONDITION2,3,4 5 digits

The Masimo MS-2013 pulse oximeter with an LNOP- Adt sensor was validated for no motion accuracy in human blood studies on healthy adult volunteers in induced hypoxia studies in the range of 70% to 100% SpO2 against a laboratory co-oximeter and ECG monitor. This variation equals plus or minus one standard deviation. Plus or minus one standard deviation encompasses 68% of the population. The Masimo MS-2013 pulse oximeter with LNOP- Adt sensor has been validated for motion accuracy in human blood studies on healthy adult volunteers in induced hypoxia studies while performing rubbing and tapping motions at 2 to 4 Hz. At an amplitude of 1 to 2 cm and non-repetitive motion between 1 to 5 Hz at an amplitude of 2 to 3cm in induced hypoxia studies in the range of 70% to 100% SpO2 against a laboratory co-oximeter and ECG monitor. This variation equals plus or minus one standard deviation. Plus or minus one standard deviation encompasses 68% of the population. The Masimo MS-2013 pulse oximeter with LNOP-Neo and Neo Pt sensors has been validated for neonatal motion accuracy in human blood studies on neonates while moving the neonates foot at 2 to 4 Hz at an amplitude of 1 to 2 cm against a laboratory co-oximeter and ECG monitor. This variation equals plus or minus one standard deviation. Plus or minus one standard deviation encompasses 68% of the population. The Masimo MS-2013 pulse oximeter has been validated for low perfusion accuracy in bench top testing against a Biotek Index 2 simulator and Masimos simulator with signal strengths of greater then 0.02% and transmission of greater than 5% for saturations ranging from 70 to 100%. This variation equals plus or minus one standard deviation. Plus or minus one standard deviation encompasses 68% of the population.

2.4.2.3

Alarm Setting Range


SPO2 MODULE MS-2013 SpO2 module SPO2 50 100 PR 20 250 bpm

2-8

0070-10-0702

Accutorr V Service Manual

Specifications

SpO2 Measurement

2.4.3

Nellcor NELL-3 SpO2 Performance


This section details the performance measurements of the NELL-3 pulse oximeter (integrated OxiMax measurement technique), which is designed in accordance with the requirement of the interface of the Nellcor NELL-3 board. The SHIP (Standard Host Interface Protocol) mode is used.

2.4.3.1

SpO2
Compatible Sensors: See Section 6.6 for a list of compatible sensors Measurement Range and Accuracy:

ACCURACY1,2 SENSOR MAX-A, MAX-P, and MAX-I OxiCliq A, OxiCliq N, OxiCliq P and OxiCliq I D-YS, DS-100A, OXI-A/N and OXI-P/I MAX-R, D-YSE and D-YSPD 1 70% 100% 2% 2.5% 3% 3.5% 0% 69% Undefined Undefined Undefined Undefined

When sensors are used on neonatal subjects as recommended, the specified accuracy range increases by 1 digit, to account for the theoretical effect on oximeter measurements of fetal hemoglobin in neonatal blood. Neonatal accuracy specifications are based upon testing neonatal sensors on healthy adult volunteers in induced hypoxia studies, in the range of 70-100% SpO2. The specified accuracy also takes into account published literature which predicts that there may be a small difference in % SpO2 reported by the oximeter when measurements from adult and fetal blood with 100% fetal hemoglobin are compared. Fetal hemoglobin is present in concentrations varying from 10% to 90% in neonatal blood, and this percentage declines over time. As the percentage of fetal hemoglobin in neonatal blood declines, the theoretical effect on accuracy due to this source is reduced

2.4.3.2

Pulse Rate
MEASUREMENT RANGE 20 250 BPM 251 300 BPM ACCURACY 3 BPM Undefined

2.4.3.3

Alarm Setting Range


SPO2 MODULE Nell-3 SpO2 module SPO2 50 100 PULSE RATE 20 250 bpm

Accutorr V Service Manual

0070-10-0702

2-9

Temperature

Specifications

2.5

Temperature
MODE monitor MEASUREMENT RANGE 25 C 44 C (77 F 111.2 F)

ACCURACY 25 C (77 F) 32 C (89.6 F), not including 32 C (89.6 F) 32 C (89.6 F) 44 C (111.2 F), including 32 C (89.6 F) 0.2 C (0.4 F)

TYPICAL MEASUREMENT TIME <60 s

0.1 C (0.2 F)

predict

35 C 43 C (95 F 109.4 F)

Not defined

typically 10 s 12 s from the moment when temperature ---- displays dynamically

2.6

Display Area, Indicator, and Controller


On the front panel of the Accutorr V, there are a panel display area, a red/yellow LED group (on the top of the device), a set of buttons, and patient cable connectors. Refer to Chapter 2 of the Accutorr V Operating Instructions (P/N 0070-10-0699-XX).

2.6.1

Display Area
The display area includes 1 LCD, 7 groups of 7-segment digit display, and 7 groups of monochrome LED indicators.

2.6.1.1

LCD

LCD TYPE Monochrome FSTN

SIZE AND RESOLUTION Matrix: 320160 dot pitch: 0.240.24

DISPLAYED CONTENTS Startup screen, menus/ dialogs, time and date, prompt messages, patient ID, patient size, trend data (including scroll bar), SpO2 waveform, and time to the next NIBP measurement Startup screen, menus/ dialogs, time and date, prompt messages, patient ID, patient size, trend data (including scroll bar), SpO2 waveform, and time to the next NIBP measurement

NOTE The brightness and contrast of the LCD are adjustable.

Color TFT

Matrix: 480272 dot pitch: 0.161250.16125

Only brightness of LCD is adjustable.

2 - 10

0070-10-0702

Accutorr V Service Manual

Specifications

Display Area, Indicator, and Controller

2.6.1.2

7-Segment Digit Display


Seven groups of 7-segment digit displays display the systolic pressure (red LED), diastolic pressure (red LED), mean pressure/cuff pressure (red LED), SpO2 (green LED), PR (red LED), temperature (red LED), and NIBP interval (red LED).

2.6.1.3

Monochrome LED Indicator


Seven groups of monochrome LED indicators display or indicate the pulse strength (green LED), ADULTPEDIATRICNEO (patient type)(green LED), Pulse Rate Source (SpO2 or NIBP)(green LED), F/C (unit of temperature)(red LED), silence function (yellow LED), NIBP measurement status (green LED), and temperature measurement type (red LED).

2.6.1.4

Power Indicator
AC power indicator Green LED Illuminates when the Accutorr V is connected to the AC mains. Battery indicator Green LED Illuminates when the unit is on and the battery is inserted. Flashes when the battery voltage is too low. On/Standby indicator Green LED Illuminates when the device is powered on or in standby mode. Darkens when the device is shut down.

2.6.2

Overlay and labeling


The front panel overlays are displayed in English and extended to other languages. The labeling is displayed only in English. Startup screen, menu fonts, and LCD prompting characters can be changed through a software upgrade. The default start logo is Accutorr V. The format of the date can be set in the maintenance mode: Chinese style (year - month day), European style (day - month - year), American style (year - day - month). Display of NIBP data: The LCD display order is: systolic pressure data, diastolic pressure data, and mean pressure data. For legibility, a / separates the data of the three pressures. Data displayed by LED digit displays: The systolic pressure data and diastolic pressure data (in the same fonts) are displayed in parallel. The font of the mean pressure data is smaller than the systolic pressure data and diastolic pressure data, and displays next to them.

Accutorr V Service Manual

0070-10-0702

2 - 11

Display Area, Indicator, and Controller

Specifications

2.6.3

Alarm Lamp
The monitor provides 2 alarm priority levels: high level alarms or low level alarms. The Alarm lamp is on the front panel of the monitor. It is a group of (red/yellow) LED indicators. During a high alarm, the red LED, of which the duty cycle is 50%, flashes at a frequency of 1.4 2.8 Hz (84 times per minute). During a low alarm, the yellow LED illuminates continuously. In case a high alarm and a low alarm occur simultaneously, only the red LED flashes.
NOTE: The audio alarm indication complies with EN60601-1-8.

2 - 12

0070-10-0702

Accutorr V Service Manual

Specifications

Display Area, Indicator, and Controller

2.6.4

Keys
There are 14 keys on the front panel of the monitor.

KEYS

DESCRIPTION Turn the monitor on, off, or switch to standby. In the operating state, press and hold for less than 1 second to switch the device to the standby or press and hold for 2s to turn the device off. Press to start an NIBP measurement. During an NIBP measurement, stops the measurement. Also, suspends interval mode until the next time NIBP Start is pressed. Start recording trend data/real-time waveform. Press again to stop recording. When recording trend data, press and hold for less than 1 second to record the current data displaying on the LCD, or press and hold for 2 seconds to record all data of the current patient. This button has dual functions: Press and hold for less than 1 second to pause an audio alarm (parameters in alarm continue to flash). Press and hold for 2 seconds to silence an alarm. Sets alarm parameters. Sets the measurement mode and time interval. Option: STAT mode, Off, and interval mode (options: 1, 2, 3, 5,10, 15, 20, 30, 60, 120, and 240 minute intervals). In the Off mode, perform measurement manually. Press to view the menu options. Press to switch between the normal views, PLETH waveform. Press to automatically create a patient ID. Press to change patient size Press to perform menu operations on the LCD.

On/Standby indicator Start NIBP Deflate

Print

Silence

Set Alarms

Interval

Menu Display Patient Info Patient Size Up Arrow OK Down Arrow

Accutorr V Service Manual

0070-10-0702

2 - 13

Audio Indicator

Specifications

2.6.5

Patient Cable Connectors


The patient cable connectors include the NIBP connector, SpO2 connector, and the optional thermometer connector.

CABLE CONNECTOR NIBP Cuff SpO2

DESCRIPTION The NIBP cuff connector label is The SpO2 connector label is SpO2. Connects the DPM, Masimo, or Nellcor cable. The Label is SpO2 with the manufacturer name. Apply standard round DPM 6 pin connector (same as used on Trio) to the monitor configured with the DPM SpO2 board. Apply the Masimo connector to the monitor configured with the MS-2013 SpO2 board. The monitor configured with the Masimo SpO2 board must be labeled with the Masimo SET registered trademark. Apply the Nellcor connector to the monitor configured with the Nell-3 SpO2 board. The monitor configured with the Nellcor SpO2 board must be labeled with the Nellcor mark. Uses optional Mindray Smartemp predictive temperature module. The connector of temperature probe (oral/axillary or rectal) is labeled T1. The oral/axillary probe is blue, and the rectal probe is red. .

Temperature probe

The Temperature probe connector label is T1.

2.7

Audio Indicator
A built-in speaker sounds audio alarms, button tones, and pulse tones. The Accutorr V supports pitch tone and multiple-level volume.

2.7.1

Pulse Tone Function


The pitch of the pulse tone is modulated according to the SpO2 value, and the frequency increases as the SpO2 value increases. The monitor can perform automatic 22-level modulation of the pulse tones pitch.

2.7.2

Multiple-Level Volume
There are 11 alarm levels. The alarm may be disabled or set to a level between 1 and 10.

2.8

Real-time Clock
Range Accuracy Resolution 0:0:0, 2001 23:59:59, 2099 At 21 3o C, 1 min/month 1 second

The real-time clock is powered by an independent battery and functions when the device is powered off.

2 - 14

0070-10-0702

Accutorr V Service Manual

Specifications

Standby Mode

2.9

Standby Mode
To enter the standby mode, the monitor must be disconnected from the patient and receive no physiological signals from a patient with no low-battery alarm. To enter the standby mode, press and confirm the operation in the pop-up dialog. After the monitor enters the standby mode: The LCD and the LEDs no longer display data. The SYS. LEDs Display flashes . The current monitoring configuration and history data are saved. All current alarms are cleared. The NURSE CALL function is disabled. The NURSE CALL function is re-enabled when the monitor resumes in normal mode. The connection with the central monitoring system stays the same as before entering the standby mode. To resume normal mode: 1. Press any key on the front panel. 2. Confirm the resume operation at the confirm dialog. After confirming the resume operation, the monitor exits the standby mode and enters the normal mode. If the operation is not confirmed within 30 seconds, the system automatically selects Cancel, and the monitor stays in the standby mode. In the following conditions, the monitor exits the standby mode and enters the normal mode: The monitor receives SpO2 physiological signals for 5 seconds. The predictive temperature sensor is removed from its sheath. The monitor is powered by battery, and a low-battery alarm occurs. Once the monitor enters the normal mode: The alarm function is re-enabled. The history data, patient information, and other settings from before the monitor entered the standby mode are restored. Data communication is restored. The LCD displays. The LEDs display parameter data. Saving trend data starts.

Accutorr V Service Manual

0070-10-0702

2 - 15

Alarm Information

Specifications

2.10
2.10.1

Alarm Information
Basics
There are two types of alarms: Physiological alarms: The physiological parameters of the monitored patient exceed the specified ranges, or the physiological exceptions of the patient cannot be judged according to a single physiological parameter. Technical alarms: The patient cannot be monitored normally and accurately due to faults in manual operation, technical faults, or faults in the monitor. The high priority alarm indicates only physiological alarms and a low battery alarm. The low priority alarm indicates only technical alarms, other than low battery. The Accutorr V provides functionality for selecting the upper limit and lower limit of each physiological alarm.

2.10.2

Alarm Notification
1. When an alarm occurs, the system notifies the user through the speaker, alarm indicator, relevant 7-segment LED digit displays, and LCD status bar. 2. When alarms of multiple levels occur simultaneously, the system notifies the user of the alarm with the highest level through the speaker and alarm indicator. 3. If the hospital Nurse Call system is available and connected to the monitor, the user can be notified of relevant alarms through the Nurse Call system.

2.10.3

Audio Alarm Pause, Audio Alarm Off, and Audio Alarm Silencing
The monitor provides functionality for pausing (disabling) an audio alarm for 120 seconds. After 120 seconds, the audio alarm notification functionality resumes. The user can resume the audio alarm notification at any time. An alarm causes the monitor to terminate the paused state. The monitor provides the ability to turn the audio alarm off globally. During audio alarm off, a remind signal reminds the user every minute. The monitor provides the ability to silence an alarm. The monitor automatically exits the alarm silenced status if a new technical or physiological alarm occurs.

2.11

Configuration management
The monitor provides a factory configuration and a user configuration. The monitor loads the selected configuration when it is switched on. If no user configuration exists, the factory configuration is automatically loaded.

2.12

Barcode scanner
The monitor can admit a patient using the barcode scanner.

2 - 16

0070-10-0702

Accutorr V Service Manual

3.0

Troubleshooting

Introduction.................................................................................................... 3-2 Part Replacement............................................................................................ 3-2 Monitor Status Check ...................................................................................... 3-2 Technical Alarm Check.................................................................................... 3-3 Troubleshooting Guide .................................................................................... 3-4

Accutorr V Service Manual

0070-10-0702

3-1

Introduction

Troubleshooting

3.1 3.2

Introduction
This chapter details potential monitor issues with possible causes and corrective actions. For additional information on troubleshooting, contact service.

Part Replacement
Printed circuit boards (PCBs), major parts, and components in the monitor are replaceable. Isolate a PCB suspected to be defective, follow the instruction to replace it, and retest. If the trouble remains, exchange the replacement PCB with the original suspicious PCB and continue troubleshooting as directed in this chapter.

3.3

Monitor Status Check


Some troubleshooting tasks may require identifying the hardware version and status of the monitor. To view the system software version and FPGA version: 1. Press 2. Press .to display the SYSTEM SETUP dialog as shown in FIGURE 3-1. or to highlight MAINTENANCE.

FIGURE 3-1 3.

SYSTEM SETUP Dialog to display the MAINTENANCE

Once MAINTENANCE is highlighted, press dialog as shown in FIGURE 3-2.

FIGURE 3-2 4. Press

MAINTENANCE dialog or to highlight VERSION.

3-2

0070-10-0702

Accutorr V Service Manual

Troubleshooting

Technical Alarm Check

5. Once VERSION is highlighted, press in FIGURE 3-3.

to display the VERSION dialog as shown

FIGURE 3-3 6. Press

Example VERSION dialog or to highlight TOTAL RUN TIME or MODULE VERSION. to display the data.

7. Once the selection is highlighted, press 8. Press

to exit to the MAINTENANCE dialog.

3.4

Technical Alarm Check


If a technical alarm message displays on the LCD, eliminate the technical alarm first, then troubleshoot according to the following instructions. For additional information on technical alarm messages, possible causes, and corrective actions, refer to the Accutorr V Operating Instructions.

Accutorr V Service Manual

0070-10-0702

3-3

Troubleshooting Guide

Troubleshooting

3.5
3.5.1
SYMPTOMS

Troubleshooting Guide
Power On/Off Failures
POSSIBLE CAUSE AC mains not connected or battery too low Power supply protection Cables defective or poorly connected CORRECTIVE ACTIONS Confirm that the AC mains is properly connected and that the battery capacity is sufficient. Refer to Power Supply Failures. Confirm that the cables from the power switch and the LED board to the keyboard, keyboard to the main board, and the power module to the main board are correctly connected. Confirm that the cables and connectors are not damaged.

The monitor fails to start. AC LED or battery LED does not light

Power switch and LED board defective Power module defective Main board Defective

Replace the power switch and LED board. Replace the power module. Replace the main board.

3.5.2
SYMPTOMS

Display Failure
POSSIBLE CAUSE Cables defective or poorly connected CORRECTIVE ACTIONS Backlight defective LCD defective Confirm that the cables from the display to the main board are correctly connected. Confirm that the cables and connectors are not damaged.

The display is black or blank, but the monitor still works correctly.

Replace the LCD. Replace the LCD. Update or upgrade FPGA. Confirm that the cable from the display to the main board is correctly connected. Confirm that the cables and connectors are not damaged.

Images overlapped or distorted

FPGA error Cables defective or poorly connected

3.5.3
SYMPTOMS

LED Digital Display and Indication Lamp Failure


POSSIBLE CAUSE Signal wires damaged CORRECTIVE ACTIONS LED digital display or indication lamp damaged Confirm that the LED signal wires are correctly connected. Confirm that the cables and connectors are not damaged.

LED digital display or indication lamp display abnormally.

Replace the LED digital display or indication lamp.

3-4

0070-10-0702

Accutorr V Service Manual

Troubleshooting

Troubleshooting Guide

3.5.4
SYMPTOMS

Alarm Problems
POSSIBLE CAUSE Cables defective or poorly connected CORRECTIVE ACTIONS Key and display board failure Main board failure Confirm that the cables from alarm LED board to keyboard and keyboard to main board are properly connected. Confirm that the connecting cables and connectors are not damaged.

The alarm lamp does not light or is extinguished, but alarm sounds

Replace the key and display board. Replace the main board. 1. Press to display the SYSTEM SETUP dialog as shown in FIGURE 3-1. 2. Press or to highlight MAINTENANCE.

Alarm does not sound, but alarm lamp lights properly

Audio alarm disabled

3. Once MAINTENANCE is highlighted, press to display the MAINTENANCE dialog as shown in FIGURE 3-2. 4. Press or to select USER MAINTENANCE. to

5. Once USER MAINTENANCE is highlighted, press display the USER MAINTENANCE dialog. 6. Enter the required password. 7. In the pop-up menu, set the [MIN ALARM VOL] to a value other than zero. Cable defective or poorly connected Audio program error identified in the system program Speaker failure Main board failure

Confirm that the cable between speaker and main board is properly connected. Confirm that the connecting cables and connectors are not damaged.

Upgrade system programs.

Replace the speaker. Replace the main board,

3.5.5
SYMPTOMS Keys do not work

Key Failure
POSSIBLE CAUSE Cables defective or poorly connected CORRECTIVE ACTIONS Keyboard failure Confirm that the cable between keyboard and main board is properly connected. Confirm that the connecting cables and connectors are not damaged.

Replace keyboard.

Accutorr V Service Manual

0070-10-0702

3-5

Troubleshooting Guide

Troubleshooting

3.5.6
SYMPTOMS No printout

Recorder Failures
POSSIBLE CAUSE Recorder module disabled Paper reversely installed Cable defective or poorly connected CORRECTIVE ACTION Confirm that the recorder status LED is on. If yes, reset the recorder. If no, check for other possible causes. Confirm that the cable between recorder and main board is appropriately connected. Confirm that the connecting cables and connectors are not damaged.

Re-install the paper roll.

Recorder power supply failure Recorder failure Poor print quality or paper not feeding properly Paper roll not properly installed Dirty thermal print head

Confirm that the power module outputs 5V DC and 16.8V DC correctly. Replace the recorder. Stop the recorder and re-install the paper roll. Confirm that the thermal print head and the paper roller do not contain foreign matter. Clean the thermal print head with an appropriate detergent.

Recorder failure

Replace the recorder.

3.5.7
SYMPTOMS

Interface Failures
POSSIBLE CAUSE Incorrect function settings CORRECTIVE ACTION 1. Press to display the SYSTEM SETUP dialog as shown in FIGURE 3-1. 2. Press or to highlight MAINTENANCE. to

No nurse call signals are issued

3. Once MAINTENANCE is highlighted, press display the MAINTENANCE dialog as shown in FIGURE 3-2. 4. Press or to select NURSE CALL. to

5. Once NURSE CALL is highlighted, press display the NURSE CALL dialog. 6. Select desired nurse call output. Main board failure Bar code scanner unable to be used Bar code scanner failure Serial port failure Replace the main board. Refer to Bar Code Scanner Failures. Replace the main board.

3.5.8
SYMPTOMS

Power Supply Failures


POSSIBLE CAUSE Battery failure Cable defective or poorly connected CORRECTIVE ACTION Replace battery. Power board failure Confirm that the cable between battery interface board and power module is correctly connected. Confirm that the cables and connectors are not damaged.

Fluctuating battery voltage

Replace the power board.

3-6

0070-10-0702

Accutorr V Service Manual

Troubleshooting

Troubleshooting Guide

SYMPTOMS Battery fails to be fully charged

POSSIBLE CAUSE Battery failure Cable defective or poorly connected

CORRECTIVE ACTION Replace battery. Confirm that the cable between battery interface board and the power module is properly connected. Confirm that the connecting cables and connectors are not damaged.

Power board failure Battery unable to be recharged Battery damaged

Replace the power board. Replace battery and recharge the replacement battery. If the replacement battery can be recharged, the original one fails. Confirm that the cable between battery interface board and the power module is properly connected. Confirm that the connecting cables and connectors are not damaged. Turn off the monitor then restart it. If the problem remains, disconnect the AC mains for 5 seconds and reconnect it, then restart the monitor. If the problem still remains, replace power board.

Cable defective or poorly connected

Power board failure No +3.3 V output No +5.0 V output No +12 V output NOTE: Power supply protection Power board failure

Replace the power board.

If the power module fails, it may damage other components (e.g. the monitor suddenly shuts down during the start-up). In this case, troubleshoot the power module using the procedure described in the previous table. If the parameter module, recorder, or bar code scanner malfunctions, check if the operating voltage is correct.

NOTE:

3.5.9
SYMPTOMS

Network Related Problems


POSSIBLE CAUSE No connection to LAN CORRECTIVE ACTIONS Confirm that the cables and connectors are in good condition and that the monitor is connected to the network. Confirm that the hub and switch facilities are correctly configured.

The monitor can not connect to the CMS.

Frequent dropouts and network disconnections Improper LAN cable connection Incorrect IP address configuration

Confirm that the LAN cable is connected. The LAN cable must not be longer than 50 m. Check for IP address conflict. Reconfigure IP address.

Accutorr V Service Manual

0070-10-0702

3-7

Troubleshooting Guide

Troubleshooting

3.5.10
SYMPTOMS

Software Upgrade Problems


POSSIBLE CAUSE Power failure or unintended power off during bootstrap upgrade Incorrect network connection CORRECTIVE ACTIONS Return the main board to the factory for repair.

Bootstrap upgrade fails

Program upgrade fails

Confirm that the monitor is connected to the network. Confirm that the hub or the switcher is functioning. Confirm that the net cables are of the right type and have been connected correctly.

Wrong upgrade package has been downloaded Incorrect IP address configuration

Incorrect upgrade package .mup files were used. Select package according to the program to upgrade. Configure a fixed IP address in range as specified for the monitor. Do not upgrade a program while the monitor is connected to a network.

3.5.11
SYMPTOMS

Bar Code Scanner Failures


POSSIBLE CAUSE Bar code scanner not powered on CORRECTIVE ACTIONS 1. Press to display the SYSTEM SETUP dialog as shown in FIGURE 3-1. 2. Press or to highlight MAINTENANCE. to

Bar code scanner does not work

3. Once MAINTENANCE is highlighted, press display the MAINTENANCE dialog as shown in FIGURE 3-2. 4. Press or

to select USER MAINTENANCE.

5. Once USER MAINTENANCE is highlighted, press to display the USER MAINTENANCE dialog. 6. Enter the required password. 7. Set [BARCODE POWER] to [ON] in the menu to switch on the bar code scanner. Incorrect interface cables used Loose Interface cables Bar code scanner cannot decode the target bar codes Bar code scanner not programmed correctly. Unreadable bar code Scanner and bar code at improper distance and angle Scanned characters incorrectly displayed on the monitor The monitors bar code recognition configuration does not meet the format of the target bar code. Scanner serial port output incorrectly set Replace interface cables. Confirm that the interface cables are attached securely. Re-program the bar code scanner to decode the specified bar codes. Check if bar codes of the same type can be decoded. Aim the scanner at the bar code and adjust the distance between the scanner and the bar code. Download the correct bar code recognition configuration for the monitor.

Configure the scanners serial port output according to the host serial port settings.

3-8

0070-10-0702

Accutorr V Service Manual

Troubleshooting

Troubleshooting Guide

3.5.12
SYMPTOMS

NIBP Measurement Failures


POSSIBLE CAUSE Cables detective Main board failure CORRECTIVE ACTIONS Confirm that the cables between NIBP module and the main board are in good condition. Replace the main board. Replace the NIBP module. Replace the air tubing.

Unable to start NIBP measurement Cuff inflation/deflation failure

Pump or valve damaged Air tubing occluded

3.5.13
SYMPTOMS

SpO2 Measurement Failure


POSSIBLE CAUSE SpO2 sensor failure Cables defective SpO2 module failure Main board failure CORRECTIVE ACTIONS Replace the SpO2 sensor. Confirm that the cable between the SpO2 module and the main board is in good condition. Replace the SpO2 module. Replace the main board.

Unable to measure SpO2

3.5.14
SYMPTOMS

Temperature Measurement Failures


POSSIBLE CAUSE Temperature probe damaged Cables detective Main board failure CORRECTIVE ACTIONS Replace the Temperature probe. Confirm that the cable between the Temperature module and the main board is in good condition. Replace the main board.

Unable to measure patient temperature

Accutorr V Service Manual

0070-10-0702

3-9

Troubleshooting Guide

Troubleshooting

This page intentionally left blank.

3 - 10

0070-10-0702

Accutorr V Service Manual

4.0

Repair and Disassembly

Tools ............................................................................................................. 4-2 Preparation for Disassembly............................................................................. 4-2 Disassembly................................................................................................... 4-3 Separating the Front and Rear Halves of the Monitor ......................................... 4-3 Disassembling the Front Housing Assembly ....................................................... 4-6 Removing the Main Rack Assembly................................................................... 4-9 Removing NIBP Module .................................................................................. 4-11 Removing the Main (CPU) Board ..................................................................... 4-13 Removing the Power Board .............................................................................. 4-13 Removing the Battery Connector Assembly......................................................... 4-13 Removing the Fan and Speaker ........................................................................ 4-14 Removing the Power Socket ............................................................................ 4-15 Disassembling the Temperature Module ............................................................. 4-16 Disassembling SpO2 Modules ......................................................................... 4-17 Disassembling the NIBP Module ....................................................................... 4-20

Accutorr V Service Manual

0070-10-0702

4-1

Tools

Repair and Disassembly

4.1

Tools
During disassembly and replacement, the following tools may be required: Phillips screwdrivers Wire cutters Sharp nose pliers Metric wrenches Metric nut drivers

4.2

Preparation for Disassembly


Before disassembling the monitor: stop monitoring patients, turn off the monitor, disconnect all the accessories and peripheral devices, remove the battery, and disconnect the monitor from the mains power supply.
CAUTION: Before disassembling the monitor, eliminate static charges. When disassembling the parts labeled with static-sensitive symbols, wear electrostatic discharge protection such as antistatic wristband or gloves. Follow the correct sequence to disassemble the monitor. Otherwise, the monitor may be damaged permanently. Disconnect all the cables before disassembling any parts. Take care not to damage any cables or connectors. Properly connect the cables or wires when reassembling the monitor to avoid short circuits. When assembling the monitor, select proper screws. If the wrong size screw is tightened by force, the monitor may be damaged and the screw or the part may not function as expected.

CAUTION:

4-2

0070-10-0702

Accutorr V Service Manual

Repair and Disassembly

Disassembly

4.3
4.3.1

Disassembly
Separating the Front and Rear Halves of the Monitor
1. If the monitor is not configured with an Optional Temperature module, proceed to step 2. If the monitor is configured with an Optional Temperature module: a. Remove the two (2) M36 cross-head screws (as shown in Figure 4-1).

M3x6 cross-head screw

M3x6 cross-head screw

FIGURE 4-1 Remove two (2) M3x6 cross-head screws. b. Disconnect the Temp cable connector from the PCBA as shown in Figure 4-2.

Temp cable connector

FIGURE 4-2 Disconnect the Temp cable connector.

Accutorr V Service Manual

0070-10-0702

4-3

Disassembly

Repair and Disassembly

2. Remove the four (4) M38 cross-head screws (as shown in Figure 4-3).

M3x8 cross-head screw

M3x8 cross-head screw

M3x8 cross-head screw

M3x8 cross-head screw

FIGURE 4-3 Remove four (4) M3x8 cross-head screws. 3. Release the clips (as shown in Figure 4-4) on the front and rear housings.

Clips

FIGURE 4-4 Release the clips.

4-4

0070-10-0702

Accutorr V Service Manual

Repair and Disassembly

Disassembly

4. Disconnect the cables and air tubing connecting the front and the rear housings. 5. Separate the front and the rear housings (Figure 4-5).

Cable Connector

Cable Connector

Cable Connector

Tubing Connection

Cable Connector

FIGURE 4-5 Separate the front and rear housings.


NOTE: Exercise care when separating the front and the rear housings. Be sure not to damage the cables and connectors.

Accutorr V Service Manual

0070-10-0702

4-5

Disassembly

Repair and Disassembly

4.3.2

Disassembling the Front Housing Assembly


1. Remove the six (6) M36 cross-head screws (as shown in Figure 4-6), and remove the keyboard.

M3x6 cross-head screw M3x6 cross-head screw

M3x6 cross-head screw M3x6 cross-head screw M3x6 cross-head screw M3x6 cross-head screw

Black-and-White LCD Front Housing Assembly

M3x6 cross-head screw M3x6 cross-head screw

M3x6 cross-head screw M3x6 cross-head screw M3x6 cross-head screw M3x6 cross-head screw

Color LCD Front Housing Assembly

FIGURE 4-6 Remove six (6) M3x6 cross-head screws.

4-6

0070-10-0702

Accutorr V Service Manual

Repair and Disassembly

Disassembly

2. For the black and white LCD model, remove the four (4) M38 cross-head screws (as shown in Figure 4-7). For the color LCD model, skip this step.

M3x8 cross-head screw M3x8 cross-head screw

M3x8 cross-head screw M3x8 cross-head screw

FIGURE 4-7 Remove four (4) M3x8 cross-head screws (black and white LCD model only). 3. Disconnect the flexible cables on the button board. 4. Remove the LCD. 5. Remove the two (2) M36 cross-head screws (as shown in FIGURE 4-8), and take out the SpO2 rack.

M3x6 cross-head screw

M3x6 cross-head screw

FIGURE 4-8 Remove two (2) M3x6 cross-head screws.

Accutorr V Service Manual

0070-10-0702

4-7

Disassembly

Repair and Disassembly

6. Remove the screw (as shown in FIGURE 4-9) to remove the SpO2 module from the SpO2 rack.

Screw

Screw

Screw

DPM SpO2 Module

Masimo SpO2 Module

Nellcor SpO2 Module

FIGURE 4-9 Remove the screw. 7. Remove the two (2) screws for the SpO2 signal cable (as shown in Figure 4-10). 8. Take out the SpO2 signal cable.
Screw Screw

FIGURE 4-10 Remove two (2) screws. 9. For DPM SpO2 only, unscrew the plastic nut for the DPM SpO2 signal cable (as shown in Figure 4-11).

Plastic Nut

FIGURE 4-11 For DPM SpO2 only, unscrew the plastic nut.

4-8

0070-10-0702

Accutorr V Service Manual

Repair and Disassembly

Disassembly

4.3.3

Removing the Main Rack Assembly


1. Remove the two (2) M36 cross-head screws (as shown in Figure 4-12). 2. Take out the recorder by pushing the clips inward.
M3x6 cross-head screw M3x6 cross-head screw

Clip

Clip

FIGURE 4-12 Remove two (2) M3x6 cross-head screws. 3. Unplug the two (2) cable connectors (as shown in Figure 4-13) from the recorder.

Cable Connector

Cable Connector

FIGURE 4-13 Unplug two (2) cable connectors.

Accutorr V Service Manual

0070-10-0702

4-9

Disassembly

Repair and Disassembly

4. If the device is not configured with an optional Temperature module, skip to Step 6. 5. If the device is configured with an optional Temperature module, remove the Temperature module board.

FIGURE 4-14 Remove optional temperature module board. 6. Remove the six (6) M36 cross-head screws (as shown in Figure 4-15). 7. Take out the main rack assembly.

M3x6 cross-head screw

M3x6 cross-head screw

M3x6 cross-head screw

M3x6 cross-head screw M3x6 cross-head screw

M3x6 cross-head screw

FIGURE 4-15 Remove six (6) M3x6 cross-head screws.

4 - 10

0070-10-0702

Accutorr V Service Manual

Repair and Disassembly

Disassembly

4.3.4

Removing NIBP Module


1. Unplug the NIBP module from the CPU Board (as shown in Figure 4-16).
Unplug the NIBP module

FIGURE 4-16 Unplug the NIBP module. 2. Remove the three (3) M36 cross-head screws (as shown in Figure 4-17). 3. Take out the NIBP rack.
M3x6 cross-head screw

M3x6 cross-head screw

M3x6 cross-head screw

FIGURE 4-17 Remove three (3) M3x6 cross-head screws.

Accutorr V Service Manual

0070-10-0702

4 - 11

Disassembly

Repair and Disassembly

4.3.5

Removing the Main (CPU) Board


1. Unplug the Main board power cable from the power supply as shown in FIGURE 4-18. 2. Cut the tie wrap to the speaker as shown in FIGURE 4-18.

Cable Connector

Cable Tie

FIGURE 4-18 Unplug the Main board cable connector and cut one (1) tie wrap. 3. Unplug the speaker cable (J10) from the Main board (see FIGURE 4-19). 4. Unplug the recorder cable (J8) from the Main board (see FIGURE 4-19). 5. Take out the main board by removing the four (4) M36 cross-head screws (as shown in FIGURE 4-19).

M3x6 cross-head screw

Recorder Cable Speaker Cable M3x6 cross-head screw

M3x6 cross-head screw

M3x6 cross-head screw

FIGURE 4-19 Unplug (2) two cable connectors and remove four (4) M3x6 cross-head screws.

4 - 12

0070-10-0702

Accutorr V Service Manual

Repair and Disassembly

Disassembly

4.3.6

Removing the Power Board


1. Unplug the battery connector cable. 2. Remove the power board by removing the three (3) M36 cross-head screws (as shown in Figure 4-20). 3. Unplug the remaining cable connectors (fan, AC mains connector, and Main (CPU) board) from the Power board.
M3x6 cross-head screw

M3x6 cross-head screw

M3x6 cross-head screw

FIGURE 4-20 Remove three (3) M3x6 cross-head screws.

4.3.7

Removing the Battery Connector Assembly


Remove the three (3) M36 cross-head screws (as shown in Figure 4-21).

M3x6 cross-head screw

M3x6 cross-head screw

M3x6 cross-head screw

FIGURE 4-21 Remove three (3) M3x6 cross-head screws.

Accutorr V Service Manual

0070-10-0702

4 - 13

Disassembly

Repair and Disassembly

4.3.8

Removing the Fan and Speaker


1. Cut the cable ties on the fan, speaker, and the main rack. 2. Remove the fan cushion (as shown in Figure 4-22).

Fan Cushion

FIGURE 4-22 Remove the Fan Cushion. 3. Remove the three (3) M325 cross-head screws (as shown in Figure 4-23). 4. Remove the fan. 5. Remove the speaker from the main rack.

M3x25 cross-head screw M3x25 cross-head screw

M3x25 cross-head screw

FIGURE 4-23 Remove Three (3) M3x25 cross-head screws.

4 - 14

0070-10-0702

Accutorr V Service Manual

Repair and Disassembly

Disassembly

4.3.9

Removing the Power Socket


1. Remove the retaining nut securing the grounding post (as shown in Figure 4-24). 2. Remove the grounding terminal.

Retaining Nut

FIGURE 4-24 Remove the retaining nut. 3. Remove the two (2) M38 cross-head screws (as shown in Figure 4-25). 4. Remove the power socket.

M3x8 cross-head screw

M3x8 cross-head screw

FIGURE 4-25 Remove Two (2) M3x8 cross-head screws.

Accutorr V Service Manual

0070-10-0702

4 - 15

Disassembly

Repair and Disassembly

4.3.10

Disassembling the Temperature Module


1. Remove the two (2) M36 cross-head screws (as shown in Figure 4-26). 2. Take the cover plate off.

M3x6 cross-head screw

M3x6 cross-head screw

FIGURE 4-26 Remove Two (2) M3x6 cross-head screws 3. Remove the three (3) M36 cross-head screws (as shown in Figure 4-27). 4. Remove the Temperature module board. 5. Remove the mylar sleeve.

M3x6 cross-head screw M3x6 cross-head screw

M3x6 cross-head screw

FIGURE 4-27 Remove three (3) M3x6 cross-head screws.

4 - 16

0070-10-0702

Accutorr V Service Manual

Repair and Disassembly

Disassembly

4.3.11 4.3.11.1

Disassembling SpO2 Modules Disassembling a DPM SpO2 module


1. Remove the M36 cross-head screw (as shown in Figure 4-28).

M3x6 cross-head screw

FIGURE 4-28 Remove one M3 X 6 cross-head screw.z 2. Remove the SpO2 board (as shown in Figure 4-29).

FIGURE 4-29 DPM SpO2 board and isolated power supply.

Accutorr V Service Manual

0070-10-0702

4 - 17

Disassembly

Repair and Disassembly

4.3.11.2

Disassembling a Masimo SpO2 module


1. Remove the isolation power board by removing the three (3) M3 nuts (as shown in Figure 4-30). 2. Remove the three (3) M36 cross-head screws.
M3 Nut M3x6 cross-head screw

M3 Nut M3x6 cross-head screw M3 Nut M3x6 cross-head screw

FIGURE 4-30 Remove three (3) M3 nuts, and three (3) M3x6 cross-head screws. 3. Remove the SpO2 board.

FIGURE 4-31 Masimo SpO2 board and isolated power supply.

4 - 18

0070-10-0702

Accutorr V Service Manual

Repair and Disassembly

Disassembly

FIGURE 4-32 Masimo isolated power supply.

4.3.11.3

Disassembling a Nellcor SpO2 module


1. Remove the M36 cross-head screw (as shown in FIGURE 4-33) to separate the SpO2 module from its isolated power supply (as shown in FIGURE 4-34).

M3x6 cross-head screw

FIGURE 4-33 Remove one (1) M3x6 cross-head screw.

FIGURE 4-34 Nellcor SpO2 board and isolated power supply.

Accutorr V Service Manual

0070-10-0702

4 - 19

Disassembly

Repair and Disassembly

4.3.12 4.3.12.1

Disassembling the NIBP Module Removing the NIBP Pump


1. Unplug the pump cable connector from the NIBP board. 2. Disconnect the tubing. 3. Cut the two tie wraps holding the NIBP pump as shown in FIGURE 4-35. 4. Remove the pump.
Pump Tubing

Tie Wraps

Pump Cable Connector

FIGURE 4-35 Cut two (2) tie wraps.

4.3.12.2

Removing the NIBP Dump Valve


1. Unplug the dump valve cable connector from the NIBP board. 2. Disconnect the tubing. 3. Remove the two (2) screws on the bottom of the NIBP rack as shown in FIGURE 4-36. 4. Remove the dump valve.

Screw

Screw

FIGURE 4-36 Remove two (2) screws.

4 - 20

0070-10-0702

Accutorr V Service Manual

Repair and Disassembly

Disassembly

4.3.12.3

Removing the NIBP Bleed Valve


1. Unplug the bleed valve cable connector from the NIBP board. 2. Disconnect the tubing. 3. Remove the two (2) screws as shown in FIGURE 4-37. 4. Remove the bleed valve.

Screws

FIGURE 4-37 Remove two (2) screws.

Accutorr V Service Manual

0070-10-0702

4 - 21

Disassembly

Repair and Disassembly

This page intentionally left blank.

4 - 22

0070-10-0702

Accutorr V Service Manual

5.0

Parts

Main Assembly .............................................................................................. 5-2 Front Assembly (Color LCD) ............................................................................. 5-3 Rear Housing Assembly................................................................................... 5-12 Main Bracket Assembly ................................................................................... 5-13 NIBP Assembly (Pump M6Q-100003- - -)........................................................... 5-16 NIBP Assembly (Pump 082-000056-00) ............................................................ 5-19 Temperature Module Subassembly .................................................................... 5-22 Temperature Power Module Subassembly .......................................................... 5-24

Accutorr V Service Manual

0070-10-0702

5-1

5.1

Main Assembly
5 6

5-2 0070-10-0702 Accutorr V Service Manual

Main Assembly

10

4 3 2 1

ITEM 1 2 3 4 5 6 7 8 9 10

PART NUMBER 801-6101-00003-00 801-6101-00023-00 M6G-020015- - M04-004012- - 6101-30-46616 801-6101-00011-00 M04-000605- - 6101-20-46742 9211-20-87369 801-6101-00010-00 M04-002505- - -

DESCRIPTION Front Shell Assembly (B&W) Front Shell Assembly (Color) Hose Silicone,1.6mmODX0.8mmID 10m/per Screw, Crosshead M36 Main Bracket Assembly Recorder module package Screw, Pan Head Phillips M3X8 Rear Overlay (6101) AC Inlet Retainer Clip Rear Housing Assembly Screw M36

Parts

FIGURE 5-1 Main Assembly

5.2

Front Assembly (Color LCD)


12 13 14

Accutorr V Service Manual 0070-10-0702 5-3

Parts

15

11

16

ITEM
17

PART NUMBER 115-016544-00 * 0010-20-12194 043-001894-00 6101-20-46722 6101-20-46723 6101-20-46762 M04-051003- - M04-021000- - 6101-20-46743

18

11 12 13 14 15 16 17 18

DESCRIPTION Front Assembly Service Kit (Color LCD) Includes all parts below: Machine Pneumatic Socket Subassembly Front Bezel (Color LCD) Alarm Light Cover (6101) Silicon Keypad (6101) Key Pad Fixer 2 Screw, Pan Head Phillips PT2.0X6 Flat Washer Key Pad Fixer (6101)

Front Assembly (Color LCD)

FIGURE 5-2 Front Bezel Assembly

NOTE:

* To upgrade the Black & White LCD requires FRU# 801-6101-00023-00. Included in this FRU are: FRU# 8016101-000211-00 (Main board), 009-002037-00 (LCD signal cable), 042-004637-00 (Ferrite Fixer Plate), 051-000875-00 (Display and Key panel), 049-000280-00 (Color LCD Rubber), 021-000056-00 (LCD) for the new assembly, and 115-016544-00. Please refer to the figures 5 -2, 5-4, 5-5 and 5-12.

5-4 0070-10-0702 Accutorr V Service Manual

Front Assembly (Color LCD) Parts

Accutorr V Service Manual 0070-10-0702 5-5

Parts

19 20 25 19 30

19

21

26 31 27

22

23

28 33 34 24 29 35

ITEM 19 20 21 22 23 24 25 26 27 28 29

PART NUMBER 6101-20-46734 6101-30-46622 801-0010-00002-00 801-6101-00017-00 801-0010-00018-00 6101-20-46719-51 6101-30-46621 801-6101-00016-00 801-6101-00015-00 801-0010-00019-00 6101-20-46720-51

DESCRIPTION SpO2 Mounting Bracket (6101) Nellcor SpO2 module package Nellcor SpO2 Board Nellcor Power Isolation Board Nellcor SpO2 Flex Cable Nellcor SpO2 Interface Bracket Masimo SpO2 Module Package Masimo SpO2 Board Masimo Power Isolation Board PCBA Masimo SpO2 Flex Cable Masimo SpO2 Interface Bracket

ITEM PART NUMBER DESCRIPTION 30 6101-30-46620 DPM SpO2 Module Package 31 801-9211-00108-00* DPM SpO2 Isolation Board* 32 801-9211-00107-00* DPM SpO2 Board* 33 801-6006-00029-00 DPM SpO2 Signal Cable 34 6101-20-46718-51 DPM SpO2 Interface Bracket 35 6101-20-46733 DPM SpO2 Interface Bracket Plate * Upgrading the 9006 SpO2 module to the 9008 SpO2 module requires FRU# 801-9211-00107-00, 801-9211-00108-00, 6101-20-46734.

Front Assembly (Color LCD)

FIGURE 5-3 SpO2 Assembly

5-6
36 37 38

Front Assembly (Color LCD)

0070-10-0702
39

ITEM 3 36

DESCRIPTION Screw, Crosshead M36 6101 Display and key panel PCBA (Color)* (FRU# 801-6101-00023-00 contains these parts.) 37 6101-20-46763 Ferrite Fixer Plate Isolation Sheet 38 042-004637-00* Ferrite Fixer Plate (Color LCD)* (FRU# 801-6101-00023-00 contains these parts.) 39 0000-10-10996 Spring Strip * Upgrading the Black-and-White LCD assembly to the new Color LCD assembly requires FRU# 801-6101-00021-00(Main board),009-002037-00(LCDsignal cable), 042-004637-00(Ferrite Fixer Plate), 051-000875-00(Display and Key panel), 049-000280-00(Color LCD Rubber), 021-000056-00(LCD) for the new assembly. Please refer to the figures 5 -2, 5-4, 5-5 and 5-12.

PART NUMBER M04-004012- - 801-6101-00022-00*

Accutorr V Service Manual

FIGURE 5-4 Display/Keyboard Assembly


Parts

Accutorr V Service Manual


3

Parts

40

0070-10-0702
43 42 41

ITEM 3 41 42 43 40

PART NUMBER DESCRIPTION M04-004012- - Screw Crosshead M36 6101-20-46732 Grounding Washer (6101) 049-000280-00* Color LCD Rubber* 049-000056-00* LCD panel TFT 3.5inch 272*480 3.3V* Front Overlay (6101), English 047-005559-00 Front Overlay (6101), Italy 047-005560-00 Front Overlay (6101), French 047-005561-00 Front Overlay (6101), German 047-005562-00 Front Overlay (6101), Spanish 047-005563-00 Front Overlay (6101), Portuguese 047-005564-00 Front Overlay (6101), Dutch 047-005565-00 * Upgrading the Black-and-White LCD assembly to the new Color LCD assembly requires FRU# 801-6101-00021-00 (Main board), 009-002037-00 (LCD signal cable), 042-004637-00 (Ferrite Fixer Plate), 051-000875-00 (Display and Key panel), 049-000280-00 (Color LCD Rubber), 021-000056-00 (LCD) for the new assembly. Please refer to the figures 5 -2, 5-4, 5-5 and 5-12.

Front Assembly (Color LCD)

FIGURE 5-5 Front Shell Assembly


5-7

5.3

Front Assembly (Black and White LCD)


44 13

5-8 0070-10-0702 Accutorr V Service Manual

Front Assembly (Black and White LCD)

14

11 15

16

ITEM 11 44 13 14 15 16 17 18

PART NUMBER 115-016543-00 0010-20-12194 6101-20-46710 6101-20-46722 6101-20-46723 6101-20-46762 M04-051003- - M04-021000- - 6101-20-46743

17 18

DESCRIPTION Front Assembly Service Kit (Black & White) Includes all parts below: Machine Pneumatic Socket Subassembly Front Bezel (6101) Alarm Light Cover (6101) Silicon Keypad (6101) Key Pad Fixer 2 Screw, Pan Head Phillips PT2.0X6 Flat Washer Key Pad Fixer (6101)

Parts

FIGURE 5-6 Front Bezel Assembly

Accutorr V Service Manual 0070-10-0702 5-9

Parts

19 20 25 19 30

19

21

26 32 27

22

23

28 33 34 24 29 35

ITEM 19 20 21 22 23 24 25 26

PART NUMBER 6101-20-46734 6101-30-46622 801-0010-00002-00 801-6101-00017-00 801-0010-00018-00 6101-20-46719-51 6101-30-46621 801-6101-00016-00

DESCRIPTION SpO2 Mounting Bracket (6101) Nellcor SpO2 module package Nellcor SpO2 Board Nellcor Power Isolation Board Nellcor SpO2 Flex Cable Nellcor SpO2 Interface Bracket Masimo SpO2 Module Package Masimo SpO2 Board

ITEM 27 28 29 30 32 33 34 35

PART NUMBER 801-6101-00015-00 801-0010-00019-00 6101-20-46720-51 6101-30-46620 801-9006-00001-00 801-6006-00029-00 6101-20-46718-51 6101-20-46733

DESCRIPTION Masimo Power Isolation Board PCBA Masimo SpO2 Flex Cable Masimo SpO2 Interface Bracket DPM SpO2 Module Package DPM SpO2 Board DPM SpO2 Signal Cable DPM SpO2 Interface Bracket DPM SpO2 Interface Bracket Plate

Front Assembly (Black and White LCD)

FIGURE 5-7 SpO2 Assembly

5 - 10
47 3 46 45 48

Front Assembly (Black and White LCD)

0070-10-0702
52 49 50 51

ITEM 3 45 46 47 48 49 50 51 52

PART NUMBER M04-004012- - M04-011002- - 0000-10-11179 6101-30-46609 6101-20-46752 6101-20-46763 M04-004015- - M04-021024- - 6006-20-39502

DESCRIPTION Screw, Crosshead M36 Hex Locknut with Conical Washer M3 LCD Module FSTN 320x160 LED White Backlight Display Keyboard PCBA Ferrite Fixer Plate Ferrite Fixer Plate Isolation Sheet Screw, Pan Head Phillips with Washer M3X8 Large Washer LCD Display Back Plate

Accutorr V Service Manual

FIGURE 5-8 Display/Keyboard Assembly

Parts

Accutorr V Service Manual 0070-10-0702


54

Parts

ITEM 3 53 54

53

PART NUMBER M04-004012- - 6101-20-46732 047-002718-00 6101-20-46741-52 6101-20-46741-53 6101-20-46741-54 6101-20-46741-55 6101-20-46741-56

DESCRIPTION Screw Crosshead M36 Grounding Washer (6101) Front Overlay (6101), English Front Overlay (6101), French Front Overlay (6101), German Front Overlay (6101), Spanish Front Overlay (6101), Portuguese Front Overlay (6101), Dutch

Front Assembly (Black and White LCD)

FIGURE 5-9 Front Shell Assembly


5 - 11

5.4

Rear Housing Assembly


ITEM PART NUMBER 801-6101-00010-00 DESCRIPTION Complete Rear Cover Service Kit. Includes all parts below, except 8016101-00009-00. Large Washer Foot Cushion Battery Door (6101) and Battery Door Link Flat Washer Tapping Screw PT38 Recorder Bracket (6101) Speaker Spacer Handle Cover (6101) Rear Label

5 - 12 0070-10-0702 Accutorr V Service Manual

Rear Housing Assembly

61 57

51 55 56 57 58 59 60 61

M04-021024- - DA6H-20-22831 801-6101-00009-00 M04-000802- - M04-003105- - 6101-20-46751 6006-20-39418 6101-20-46713 047-002714-00

60

59

58

55

56

57

FIGURE 5-10 Rear Housing Assembly

Parts

5.5

Main Bracket Assembly


3 63 74 75 76 77

Accutorr V Service Manual 0070-10-0702 5 - 13

Parts

72 62 71

73

78 79

80 67 64

81

65 84 66 83 82

68 69 70

ITEM 3 62 63 64 65 66 67 68

PART NUMBER M04-004012- - 6101-20-46725 6101-20-46736 6101-20-46735-51 6006-20-39451 0000-10-10996 6101-20-46738 6101-20-46753

DESCRIPTION Screw, Crosshead M36 Main Bracket (6101) Power Supply Board Isolation Sheet (6101) I/O Overlay (6101) Main Board Insulating Plate Spring Strip Battery Box Isolation Sheet (6101) Battery Hook Torsion Spring

ITEM 69 70 71 72 73 74 75 76 77 78 79 80 81 82 83 84

PART NUMBER 7000-20-24365 6006-20-39486 6101-20-46739 M04-051172- - 6101-20-46731 801-9201-00001-00 6006-20-39379 801-6006-00048-00 801-6006-00019-00 2102-20-17166 M04-004401- - M04-004504- - M04-021003- - 0509-20-00098 M04-000405- - 6006-21-39386

DESCRIPTION Torsion Spring Battery Bar Setscrew Fan Gasket (6101) Screw, Flat Head Phillips M3X25 Fan Binder Plate (6101) Fan (KDE1204PKV3) Speaker Spacer Speaker and Cables Lithium Battery Connector Assembly Protective Ground Label Hex Nut, M6 Spring Washer Flat Washer Grounding Terminal Screw, Crosshead M38 Power Board AC Input cable

Main Bracket Assembly

FIGURE 5-11 Main Bracket Assembly

5 - 14
89 85 88 45 87 90 86 91 97 3 96 95 94 93

Main Bracket Assembly

92

0070-10-0702
3

ITEM 90 91 92

ITEM 3 45 85 86 87 88 89

PART NUMBER M04-004012- - M04-011002- - 6101-30-46601 M04-000106- - 6101-20-46611 6006-20-39492 6101-30-46618

DESCRIPTION Screw, Crosshead M36 Nut, M3 with Spring Washer AC-DC Power Supply PCBA Bolt, M37 Keyboard Power Cable Recorder Power Cable NIBP Assembly

PART NUMBER DESCRIPTION 6006-20-39381 Recorder Signal Cable 6006-20-39380 Host Board Power Cable 6006-20-39384 LCD Signal Cable (for B&W LCD) 009-002037-00* LCD Signal Cable (for color LCD) * 93 M09A-20-62081 Communication Cable from Host Board to Isolation Board 94 6006-20-39382 SpO2 Module Cable 95 6006-20-39383 Keyboard Signal Cable 96 M05-010R03- - Button Battery Lithium 3V35mAh D12.5x2.0 97** 801-6101-00001-00 Main Board (for B&W LCD) 801-6101-00021-00* Main Board (for color LCD)* * Upgrading the Black-and-White LCD assembly to the new Color LCD assembly requires FRU# 801-6101-00021-00 (Main board),009-002037-00 (LCD signal cable), 042-004637-00 (Ferrite Fixer Plate), 051-000875-00 (Display and Key panel), 049-000280-00 (Color LCD Rubber), 021-000056-00 (LCD) for the new assembly. Please refer to the figures 5 -2, 5-4, 5-5 and 5-12.

Accutorr V Service Manual

Parts

FIGURE 5-12 Main Bracket Assembly

NOTE:

**After replacing the Main Board, a System Reconfiguration should be performed by following this procedure: While the monitor is powered on, press and hold the MENU key for two seconds. While the MENU key is still depressed, press and hold the power ON/OFF key. Verify that the message Clear system config and restarting appears. Release the MENU and ON/OFF keys. The monitor will shutdown and restart. Then the message Checking system config... appears. After the monitor is successfully configured, the message System configure succeed. Restarting... is displayed and the monitor shuts down and restarts automatically.

Accutorr V Service Manual 0070-10-0702 5 - 15

Parts Main Bracket Assembly

5.6

NIBP Assembly (Pump M6Q-100003- - -)


Accutorr V models with serial numbers lower than S/N A7503017K9 are configured with the old NIBP pump (083-000056-00).

5 - 16 0070-10-0702 Accutorr V Service Manual

NIBP Assembly (Pump M6Q-100003- - -)

10

ITEM 10

PART NUMBER M04-002505- - -

DESCRIPTION Screw, M36

FIGURE 5-13 NIBP Assembly (Pump M6Q-100003- - -)

Parts

5 - 17
104 102 106 103 102 105 101 01

NIBP Assembly (Pump M6Q-100003- - -)

0070-10-0702 Accutorr V Service Manual

00

100

99 98

ITEM 98 99 100 101 102 103 104 105 106

PART NUMBER M04-051004- - 801-6100-00019-00 A21-000002- - S1-0378-02-0004 S1-0008-10-0206 3001-10-07066 6101-20-46756 6101-20-46755 6101-20-46610

DESCRIPTION Screw, PT2.66 NIBP PCBA Module Tubing, Silicone,1/8"X1/4" X100ft,2800546-100 Filtered Orifice Restricto Tubing, Silicone,1/16"X3/16",TYGON 3350 (1foot) T-Connector NIBP Bracket Manifold NIBP Manifold Cable, Main Board to NIBP Module Board

FIGURE 5-14 NIBP Pneumatics (Pump M6Q-100003- - -)

Parts

Accutorr V Service Manual 0070-10-0702 5 - 18

Parts

111 114 112

111

115 110

ITEM 107 108 109 110 111 112 113 114 115

PART NUMBER 047-001582-00* M04-051137- - M90-000002-00 042-001417-00* M6Q-100003- - 6101-20-46761 082-000057-00 M04-004905- - 082-000058-00 009-000588-00 A90-000031- - -

113 114 115

DESCRIPTION NIBP Bracket Label* Screw, M24 Insulation Washer F2 NIBP Bracket* (FRU# 801-6101-00018-00 contains these parts.*) Rolling Pump Pump Cushion Valve Solenoid Screw, M318 Valve Pump connection cable Cable Tie CHS-4x150mm

NIBP Assembly (Pump M6Q-100003- - -)

* Upgrading the old pump (082-000056-00) to the new rolling pump (M6Q-100003- - -) requi FRU# 801-6101-00018-00 (pump kit), 047-001582-00 (label), and 042-001417-00 (bracke See section 5.7 (pg. 5-19) NIBP Assembly (Pump 082-000056-00) for the old pump assem

FIGURE 5-15 NIBP Mechanical (Pump 082-000056-00)

5.7

NIBP Assembly (Pump 082-000056-00)

5 - 19 0070-10-0702 Accutorr V Service Manual

NIBP Assembly (Pump 082-000056-00)

10

ITEM 10

PART NUMBER M04-002505- - -

DESCRIPTION Screw, M36

Parts

FIGURE 5-16 NIBP Assembly (Pump 082-000056-00)

5 - 20
103 102

NIBP Assembly (Pump 082-000056-00)

102

104

105

106

100

0070-10-0702 Accutorr V Service Manual

101

100

99 98

ITEM 98 99 100 101 102 103 104 105 106

PART NUMBER M04-051004- - 801-6100-00019-00 A21-000002- - S1-0378-02-0004 S1-0008-10-0206 3001-10-07066 6101-20-46756 6101-20-46755 6101-20-46610

DESCRIPTION Screw, PT2.66 NIBP PCBA Module Tubing, Silicone,1/8"X1/4" X100ft,2800546-100 Filtered Orifice Restricto Tubing, Silicone,1/16"X3/16",TYGON 3350 (1foot) T-Connector NIBP Bracket Manifold NIBP Manifold Cable, Main Board to NIBP Module Board

FIGURE 5-17 NIBP Pneumatics (Pump 082-000056-00)

Parts

Accutorr V Service Manual 0070-10-0702 5 - 21

Parts

113

114

115

112

118

117

116

108

ITEM 116 108 109 117 118 112 113 114 115

PART NUMBER 6101-20-46737 M04-051137- - M90-000002-00 6101-20-46730 082-000056-00* 6101-20-46761 801-6101-00002-00 082-000057-00 M04-004905- - 082-000058-00

109

DESCRIPTION NIBP Fram Overlay (6101) Screw, M24 Insulation Washer F2 NIBP Bracket (6101) Pump* Pump Cushion Pump Kit (includes pump 082-000056-00 and cushion 6101-20-46761) Valve Solenoid Screw, M318 Valve

NIBP Assembly (Pump 082-000056-00)

* Upgrading the old pump (082-000056-00) to the new rolling pump (M6Q-100003- - -) requires FRU# 801-6101-00018-00 (pump kit), 047-001582-00 (label), and 042-00141700 (bracket). See section 5.6 (pg. 5-16) NIBP Assembly (Pump M6Q-100003- - -) for the new rolling pump assembly.

FIGURE 5-18 NIBP Mechanical (Pump 082-000056-00)

5.8

Temperature Module Subassembly

Accutorr V Service Manual 0070-10-0702 5 - 22

Parts

10

121

120

119

ITEM 10 119 120 121

PART NUMBER M04-002505- - 6101-20-46759 801-6006-00044-00 6101-20-46716

DESCRIPTION Screw, M36 Temp Module Water Ingress Blocker Temp Measurement Board Subassembly Temp Module Inner Cover

Temperature Module Subassembly

FIGURE 5-19 Temperature Module Subassembly

5 - 23
125 124 123

Temperature Module Subassembly

0070-10-0702
122

ITEM 122 123 124 125

PART NUMBER 6101-20-46717 6101-20-46715 6101-20-46724 M04-003905- - -

DESCRIPTION Temp Probe Cover Holder Temp Module External Cover Temperature Probe Holder Ring (6101) Screw, Phillips PT36

Accutorr V Service Manual

FIGURE 5-20 Temperature Module Subassembly


Parts

5.9

Temperature Power Module Subassembly


127 126

Accutorr V Service Manual 0070-10-0702 5 - 24

Parts

128

10

Temperature Power Module Subassembly

ITEM 10 126 127 128

PART NUMBER M04-002505- - M09A-20-62082 6101-20-46745 801-6006-00043-00

DESCRIPTION Screw, M36 Communication Cable from Power Isolation Board to Temp Measurement Board Rear Inner Cover Plate (6101) Temp Module Power Supply Board PCBA

FIGURE 5-21 Temperature Power Module Subassembly

6.0

Test and Calibration

Introduction.................................................................................................... 6-2 Visual Test...................................................................................................... 6-4 Power-on Test ................................................................................................. 6-4 NIBP Calibration ............................................................................................ 6-4 NIBP Accuracy Test ......................................................................................... 6-8 SpO2 Test ...................................................................................................... 6-12 Testing the Optional Temperature Module .......................................................... 6-14 Nurse Call Performance Test ............................................................................ 6-16 Bar Code Scanner Test .................................................................................... 6-18 Electrical Safety Tests ...................................................................................... 6-20 Recorder Check.............................................................................................. 6-22 Software upgrade........................................................................................... 6-23

Accutorr V Service Manual

0070-10-0702

6-1

Introduction

Test and Calibration

6.1

Introduction
To ensure the monitor always functions properly, qualified personnel should perform regular inspections, maintenance, and testing. This chapter provides the test procedures for the monitor and includes recommended test equipment and frequency. Qualified personnel should perform the testing and maintenance procedures as required and use appropriate test equipment. The testing procedures provided in this chapter are intended to verify that the monitor meets the performance specifications. If the monitor or a module fails to perform as specified in any test, repair or replacement is required to correct the problem. If the problem persists, contact service.
CAUTION: NOTE: All tests should be performed by qualified personnel only. Qualified personnel should acquaint themselves with the test tools and confirm that test tools and cables are applicable. When fields or buttons in a dialog have a line through them, they are not available for selection or modification.

NOTE:

6-2

0070-10-0702

Accutorr V Service Manual

Test and Calibration

Introduction

6.1.1

Test Report
After completing the tests, service personnel should record test results in this table and retain them for hospital records.

TEST EQUIPMENT Name Model/PN Next Calibration Due Date

TEST RECORD No. 1 2 CONCLUSION Pass/Fail: Tested by: Date: Test Item Test Site Test Results

6.1.2
Visual test Power on test

Recommended Frequency
FREQUENCY When first installing or after reinstalling. 1. When first installing or after reinstalling. 2. Following any maintenance or replacement of any main unit parts. Accuracy test Leakage test Calibration 1. If the user suspects that the measurement is incorrect. 2. Following any repairs or replacement of the NIBP module. 3. At least once every two years.

CHECK/MAINTENANCE ITEM

NIBP tests

SpO2 test Temperature test Bar code scanner test Electrical safety tests Enclosure leakage current test Earth leakage current test Patient leakage current test Patient auxiliary current test Recorder check Following any repair or replacement of the recorder. If the user suspects that bar code scan is incorrect. 1. Following any repair or replacement of the power module. 2. At least once every two years.

Accutorr V Service Manual

0070-10-0702

6-3

Visual Test

Test and Calibration

6.2

Visual Test
Inspect the equipment for obvious signs of damage. Follow these guidelines when inspecting the equipment: Carefully inspect the case, the display screen, and the buttons for physical damage. Inspect all external connections for loose connectors, bent pins, or frayed cables. Inspect all connectors on the equipment for loose connectors or bent pins. Make sure that the labels on the equipment are clearly legible.

6.3

Power-on Test
Verify that the monitor can power up correctly by following this procedure: 1. Insert the lithium battery in the battery compartment and connect the monitor to the AC mains. The AC mains LED and battery LED light. 2. Press the Power On/Off button to turn on the monitor. The operating status LED built into the Power On/Off button illuminates. The system beeps to indicate it has passed the alarm sounds self test. All the LEDs on the front panel illuminate. The technical alarm lamp illuminates yellow, then red, and then turns off to indicate it has passed the alarm lamp self test. The start-up screen clears, and the monitor enters the main screen, indicating start-up is complete.

6.4

NIBP Calibration
Required Tools: One (1) T-Connector Three (3) pieces of Tubing One (1) Metal Vessel with volume of 500 25 ml One (1) Reference manometer with accuracy of 1 mmHg To calibrate NIBP: 1. Attach the calibration vessel and reference manometer as shown in FIGURE 6-1.

Accutorr V

Reference Manometer

Hose

Calibration Vessel

FIGURE 6-1 NIBP calibration configuration 2. If needed, press (11) to display the Normal Screen.

6-4

0070-10-0702

Accutorr V Service Manual

Test and Calibration

NIBP Calibration

3. Press 4. Press

(12) to display the SYSTEM SETUP dialog. (16) or (19) to highlight MAINTENANCE.

5. Once MAINTENANCE is highlighted, press (18) to display the MAINTENANCE dialog as shown in FIGURE 6-2.

FIGURE 6-2 MAINTENANCE dialog 6. Press (16) or (19) to highlight NIBP TOOLS. (18) to display the NIBP TOOLS

7. Once NIBP TOOLS is highlighted, press dialog. 8. Press (16) or

(19) to highlight CALIBRATION, as shown in FIGURE 6-3.

FIGURE 6-3 NIBP TOOLS dialog 9. Once CALIBRATION is highlighted, press dialog, as shown in FIGURE 6-4. (18) to display the Enter Password

FIGURE 6-4 ENTER PASSWORD dialog

Accutorr V Service Manual

0070-10-0702

6-5

NIBP Calibration

Test and Calibration

10. In the Enter Password dialog, press 11. Press 12. Press 13. Press 14. Press 15. Press 16. Press 17. Press 18. Press 19. Press 20. Press 21. Press 22. Press
NOTE:

(18). See FIGURE 6-4.

(16) or (18). (16) or (18). (16) or (18). (16) or (18). (16) or (18). (16) or (18).

(19) to set the first password digit to 3. (19) to highlight the second password digit. (19) to set the second password digit to 2. (19) to highlight the third password digit. (19) to set the third password digit to 1. (19) to highlight OK.

When the NIBP CALIBRATION dialog is first displayed, INFLATE is highlighted, as shown in FIGURE 6-5.

FIGURE 6-5 NIBP CALIBRATION dialog with INFLATE highlighted 23. Press
NOTE:

(18) to start pressurizing the calibration vessel.


Use (16) or (19) to highlight CANCEL, then press

(18) to cancel the calibration. NOTE: After the calibration vessel is pressurized, the reference pressure field is highlighted, as shown in FIGURE 6-6.

FIGURE 6-6 NIBP CALIBRATION dialog with the reference pressure field highlighted

6-6

0070-10-0702

Accutorr V Service Manual

Test and Calibration

NIBP Calibration

24. Press the (16) or (19) to adjust the reference pressure value in 1 mm increments until it matches the MEASURED PRESSURE value. 25. Press (18) to set the reference pressure value. After the reference pressure value is set, ACCEPT is highlighted, as shown in FIGURE 6-6.

FIGURE 6-7 NIBP CALIBRATION dialog with ACCEPT highlighted 26. Press the 27. Press
NOTE:

(16) or

(19) to select ACCEPT.

(18) to accept the calibration.


Use (16) or (19) to highlight CANCEL, then press

(18) to cancel the calibration.

The Accutorr V uses the new calibration factor until another calibration sequence is performed.

Accutorr V Service Manual

0070-10-0702

6-7

NIBP Accuracy Test

Test and Calibration

6.5

NIBP Accuracy Test


Required Tools: Two (2) T-Connectors Five (5) pieces of tubing One (1) Metal Vessel with volume of 500 25 ml One (1) Reference manometer with accuracy of 1 mmHg One (1) Ball Pump To calibrate NIBP: 1. Attach the calibration vessel, ball pump, and reference manometer as shown in FIGURE 6-8.

Accutorr V

Hose

Reference Manometer

Calibration Vessel

Ball Pump

FIGURE 6-8 NIBP accuracy test configuration 2. If needed, press 3. Press 4. Press or to display the Normal Screen. to highlight MAINTENANCE. to display the MAINTENANCE

to display the SYSTEM SETUP dialog.

5. Once MAINTENANCE is highlighted, press dialog. 6. Press or to highlight NIBP TOOLS. 7. Once NIBP TOOLS is highlighted, press shown in FIGURE 6-3.

to display the NIBP TOOLS dialog as

6-8

0070-10-0702

Accutorr V Service Manual

Test and Calibration

NIBP Accuracy Test

8. Press

or

to highlight ACCURACY TEST. to start the ACCURACY TEST as

9. Once ACCURACY TEST is highlighted, press shown in FIGURE 6-9.

FIGURE 6-9 NIBP TOOLS dialog running the accuracy test


NOTE: The ACCURACY TEST button changes to STOP ACCURACY TEST. Use then press or to highlight STOP ACCURACY TEST, to cancel the accuracy test.

10. Use the ball pump to pressurize the vessel to each of the three different ranges shown in following table.
RANGE Low approximately 50 mmHg Medium approximately 150 mmHg High approximately 250 mmHg ACCURACY 3 mmHg 3 mmHg 3 mmHg

The pressure value in the MAP LED display should match the value shown on the reference manometer within the accuracy for each test range as shown in the table. 11. Use or accuracy test. to highlight STOP ACCURACY TEST, then press to end the

If the accuracy test failed, send the Accutorr V to service for repair.

6.5.1

NIBP Leakage Test


Required Tools: One (1) piece of tubing One (1) Cylinder One (1) Metal Vessel with volume of 500 25 ml To perform the NIBP leakage test: 1. Set Patient size to adult by pressing changes with each key press. until Adult is selected. The patient size

Accutorr V Service Manual

0070-10-0702

6-9

NIBP Accuracy Test

Test and Calibration

Adult

Pediatric

Neonate

FIGURE 6-10 Patient size graphics and indicators The Patient size indicator illuminates to indicate the selected size as shown in FIGURE 610. The factory default setting for the Patient size is Adult. 2. Connect the metal vessel with the NIBP connector on the Accutorr V monitor.

Accutorr V

Hose

Calibration Vessel

FIGURE 6-11 Leakage test configuration 3. If needed, press 4. Press 5. Press or to display the Normal Screen. to highlight MAINTENANCE. to display the MAINTENANCE

to display the SYSTEM SETUP dialog.

6. Once MAINTENANCE is highlighted, press dialog. 7. Press or to highlight NIBP TOOLS. 8. Once NIBP TOOLS is highlighted, press shown in FIGURE 6-3. 9. Press or to highlight LEAK TEST.

to display the NIBP TOOLS dialog as

10. Once LEAK TEST is highlighted, press to start the leak test. The Accutorr V automatically deflates in approximately 60s, and the NIBP leakage test is complete.

FIGURE 6-12 NIBP TOOLS dialog running the leak test


NOTE: The LEAK TEST button changes to STOP LEAK TEST. Use or to highlight STOP LEAK TEST, then press to cancel the leak test.

6 - 10

0070-10-0702

Accutorr V Service Manual

Test and Calibration

NIBP Accuracy Test

11. If the system leaks, the message PNEUMATIC LEAK will be displayed in the technical alarm area/prompt area. In this case, check for a loose connection and perform the test again.
NOTE: If the system does not leak, the Accutorr V does not display a message.

Accutorr V Service Manual

0070-10-0702

6 - 11

SpO2 Test

Test and Calibration

6.6

SpO2 Test
NOTE: A functional tester cannot be used to assess the accuracy of a pulse oximeter probe or a pulse oximeter monitor.

Required Tool: SpO2 simulator. For the monitor equipped with the DPM SpO2 module, Nellcor SpO2 module, or Masimo SpO2 module, a calibrated SpO2 simulator is recommended.

6.6.1

SpO2 Test Under Normal Conditions


1. Connect the SpO2 simulator to the SpO2 sensor. 2. Select the model and the manufacturer of the SpO2 module under test, and then configure the SpO2 simulator as follows: SpO2 96%; PR 80 bpm. 3. The displayed SpO2 and PR values should be within the ranges listed below.

6.6.2

SpO2 Test in Motion Mode


1. Connect the SpO2 simulator to the SpO2 sensor. 2. Select the model and the manufacturer of the SpO2 module under test; take measurement in the motion mode preset by the SpO2 simulator. 3. The displayed SpO2 and PR values should be within the ranges listed below.

MANUFACTURER DPM Compatible Sensors Masimo Compatible Sensors

SPO2 SENSOR 512E, 512G, 512F, 512H, 518B, 520A, 520P, 520I 520N LNCS-NeoPt-L, LNCS Neo-L, LNOP DCI, LNOP DCIP, LNOP TCI, LNOP YI-Multisite, LNOP DCSC, LNOP Adt, LNOP Pdt, LNOP II Inf-L, LNOP II NeoNeonatal L, LNOP NeoPt Preterm Neonatal Y, LNOP II NeoPt Preterm Neonatal Y LNCS Inf-L, LNCS-Pdtx, LNCS-Adtx, LNCS DCI,LNCS DC-I Pt, TC-I

SPO2 96%2% 96%3% 96%3% (without motion) 96%3% (with motion) 96%2% (without motion) 96%3% (with motion) 96%2% 96%3% 96%4%

PR (BPM) 803

803 bpm (without motion) 805 bpm (with motion)

Nellcor Compatible Sensors

MAX-A, MAX-P, MAX-I, DS-100A, OXI-A/N (Adult), OXI-P/I, MAX-N OXI-A/N (Neonate)

803 bpm

NOTE:

The SpO2 simulator can only be used to verify that the pulse oximeter operates properly. It cannot be used to verify the accuracy of the pulse oximeter or the SpO2 sensor. To verify the accuracy, clinical tests are required.

Contact Technical Support if the SpO2 test fails.

6 - 12

0070-10-0702

Accutorr V Service Manual

Test and Calibration

SpO2 Test

6.6.3 6.6.3.1

Summary of Test Methods DPM SpO2


Measurement validation: The DPM SpO2 module accuracy has been validated in human studies against arterial blood sample reference measured with a CO-oximeter. Pulse oximeter measurements are statistically distributed, and about two-thirds of the measurements can be expected to fall within the specified accuracy compared to CO-oximeter measurements.

6.6.3.2 6.6.3.3

Nellcor SpO2
The Nellcor SpO2 module was tested for accuracy using a simulator.

Masimo SpO2
The Masimo SpO2 module was tested by Masimo to verify accuracy.

Accutorr V Service Manual

0070-10-0702

6 - 13

Testing the Optional Temperature Module

Test and Calibration

6.7

Testing the Optional Temperature Module


Required Tool: Thermostatic oil tank, HART 7102 recommended 1. Set the temperature of the oil tank to 37 C and conduct the test after the temperature stabilizes. 2. Press to display the SYSTEM SETUP dialog as shown in FIGURE 6-13.

FIGURE 6-13 SYSTEM SETUP dialog 3. Press or to highlight TEMP SETUP. to display the TEMP SETUP dialog as

4. Once TEMP SETUP is highlighted, press shown in FIGURE 6-14.

FIGURE 6-14 Example TEMP SETUP dialog 5. Press 7. Press 9. Press or or or to highlight the TEMP TYPE pull-down list. to view the selections. to set it. to view the selections. to change the selection to MONITOR. to highlight the TEMP UNIT pull-down list.

6. Once the pull-down list is highlighted, press 8. Once MONITOR is highlighted, press 10. Once the pull-down list is highlighted, press
NOTE:

For this test, the monitor and test tank temperature units are in Celsius.

11. Press

or

to change the selection to C. to set it. to exit the TEMP SETUP dialog.

12. Once C is highlighted, press 13. Once the choices are set, press

6 - 14

0070-10-0702

Accutorr V Service Manual

Test and Calibration

Testing the Optional Temperature Module

14. Remove the Temperature probe from the probe sheath, insert a probe cover, and place the probe into the oil tank. 15. Wait till the Temp value displayed on the monitor stabilizes. Verify that the displayed value is 370.1 C. Contact Technical Support if the temperature test fails.

Accutorr V Service Manual

0070-10-0702

6 - 15

Nurse Call Performance Test

Test and Calibration

6.8

Nurse Call Performance Test


Required Tool: Multimeter 1. Connect the nurse call cable to the analog output connector. 2. If needed, press 3. Press 4. Press or to display the Normal Screen. to highlight MAINTENANCE. to display the MAINTENANCE to display the SYSTEM SETUP dialog.

5. Once MAINTENANCE is highlighted, press dialog shown in FIGURE 6-2. 6. Press or to highlight NURSE CALL. 7. Once NURSE CALL is highlighted, press dialog shown in FIGURE 6-15.

to display the NURSE CALL SETUP

FIGURE 6-15 NURSE CALL SETUP dialog 8. Press 10. Press 12. Press 14. Press 16. Press 18. Press or or or or or or to highlight the SIGNAL DURATION pull-down list. to select it. to highlight the PULSE. to select it. to select it. to highlight the SIGNAL TYPE pull-down list. to highlight the NORMAL OPEN. to select it. to highlight either HIGH or LOW for ALM LEV. to select it. to select it. to highlight either TECH or PHYS for ALM TYPE.

9. Once the SIGNAL DURATION pull-down list is highlighted, press 11. Once the PULSE is highlighted, press

13. Once the SIGNAL TYPE pull-down list is highlighted, press 15. Once the NORMAL OPEN is highlighted, press 17. Once an ALM LEV is highlighted, press 19. Once an ALM TYPE is highlighted, press

20. Trigger an alarm and measure the contact output with the multimeter. The output should be square waves with an interval of 1s. 21. Press 23. Press or or to highlight the SIGNAL DURATION pull-down list. to select it. to highlight the CONTINUOUS. to select it. 22. Once the SIGNAL DURATION pull-down list is highlighted, press 24. Once the CONTINUOUS is highlighted, press

6 - 16

0070-10-0702

Accutorr V Service Manual

Test and Calibration

Nurse Call Performance Test

25. In the NURSE CALL SETUP dialog, set SIGNAL TYPE to NORMAL OPEN. Trigger an alarm and measure the contact output with the multimeter. The output should be continuous high level. Contact Technical Support if the nurse call test fails.

Accutorr V Service Manual

0070-10-0702

6 - 17

Bar Code Scanner Test

Test and Calibration

6.9

Bar Code Scanner Test


Required Tool: None 1. Connect the barcode reader to the RS-232 connector on the back of the Accutorr V. 2. Press 3. Press to display the SYSTEM SETUP dialog FIGURE 6-13. or to select MAINTENANCE. to display the MAINTENANCE

4. Once MAINTENANCE is highlighted, press dialog as shown in FIGURE 6-16.

FIGURE 6-16 MAINTENANCE dialog 5. Press or to select USER MAINTENANCE. to display the Enter Password

6. Once USER MAINTENANCE is highlighted, press dialog, as shown in FIGURE 6-17.

FIGURE 6-17 ENTER PASSWORD dialog

6 - 18

0070-10-0702

Accutorr V Service Manual

Test and Calibration

Bar Code Scanner Test

7. In the Enter Password dialog, press

8. Enter the password 321 to display the USER MAINTENANCE dialog as shown in FIGURE 6-18. See Section 6.4, steps 10 22 for the password procedure.

FIGURE 6-18 Example USER MAINTENANCE dialog 9. Press 11. Press or or to highlight the BARCODE POWER selection field. select it. select it. to select ON.

10. Once the selection field is highlighted, press 12. Once the selection is highlighted, press
NOTE:

When the RS-232 connector is used for DIAP, barcode power must be set to OFF.

13. Aim the bar code scanner at the target bar code. Adjust the field of view to capture the bar code. 14. Hold the trigger until the bar code scanner beeps, indicating the bar code is successfully decoded. The indication lamp turns green and scanned characters are displayed on the monitor. Contact Technical Support if the bar code scanner does not work as described.

Accutorr V Service Manual

0070-10-0702

6 - 19

Electrical Safety Tests

Test and Calibration

6.10

Electrical Safety Tests


WARNING: Electrical safety tests are a proven means of verifying the electrical safety of the monitor. They are intended for determining potential electrical hazards. Failure to identify these hazards in a timely manner may cause personnel injury. WARNING: Commercially available test equipment such as a safety analyzer can be used for electrical safety tests. Verify that the test equipment can be safely and reliably used with the monitor before use. The service personnel should acquaint themselves with the use of the test equipment. WARNING: Electrical safety tests should meet the requirements of the latest editions of EN 60601-1 and UL 60601. WARNING: These electrical safety tests do not supersede local requirements. WARNING: All devices using the AC mains and connected to medical equipment within patient environments must meet the requirements of the IEC 60601-1-1 medical electrical systems standard and should be put under electrical safety tests at the frequency recommended for the monitor.

Electrical safety tests are intended to check the potential electrical hazards to the patient, operator, or service personnel. Electrical safety tests should be performed under normal ambient conditions of temperature, humidity, and atmospheric pressure. The electrical safety test plan described here uses a calibrated safety analyzer. Different safety analyzers may be used. Choose an applicable calibrated safety analyzer and test plan.

6.10.1

Enclosure Leakage Current Test


1. Connect the safety analyzer to an AC power supply. 2. Connect SUM terminal of the applied part connection apparatus to RA input terminal of safety analyzer, another terminal to the applied part of EUT. 3. Connect the EUT to the safety analyzers auxiliary output connector by using a power cord. 4. Attach on end of the red lead to the red input terminal of the analyzer, and the other end to tinsel over the enclosure of the EUT. 5. Power on the safety analyzer and then press the Enclosure leakage button on the analyzers panel to enter the enclosure leakage test screen. Under normal condition, the enclosure leakage current should be no greater than 100A. Under single fault condition, the leakage current should be no greater than 300A.

6.10.2

Earth Leakage Current Test


1. Connect the safety analyzer to an AC power supply. 2. Connect the SUM terminal of the applied part connection apparatus to the RA input terminal of the safety analyzer, another terminal to the applied part of EUT.

6 - 20

0070-10-0702

Accutorr V Service Manual

Test and Calibration

Electrical Safety Tests

3. Connect the EUT to the safety analyzers auxiliary output connector by using a power cord. 4. Power on the safety analyzer and then press the Earth leakage button on the analyzers panel to enter the earth leakage test screen. Under normal condition, the earth leakage current should be no greater than 300A. Under single fault condition, the leakage current should be no greater than 1000 A.

6.10.3

Patient Leakage Current Test


1. Connect the safety analyzer to an AC source. 2. Connect the SUM terminal of the applied part connection apparatus to RA input terminal of safety analyzer, another terminal to the applied part of EUT. 3. Connect the EUT to the safety analyzers auxiliary output connector by using a power cord. 4. Power on the safety analyzer and then press the Patient leakage on the safety analyzers panel. 5. Repeatedly press the APPLIED PART button to measure AC and DC leakage alternatively. DC leakage reading is followed by DC. Under normal status, the patient leakage current should be no greater than 100A AC, 10A DC. Under single fault condition, the leakage current should be no greater than 500A AC, 50A DC.

6.10.4

Patient Auxiliary Leakage Current Test


1. Connect the safety analyzer to an AC source. 2. Connect the equipment under test (EUT) to the analyzers auxiliary output connector by using a power cord. 3. Connect the sensors of the applied part to the applied part connection apparatus, whose RA-P terminal is connected to safety analyzers RA terminal and SUM terminal to safety analyzers LA terminal. RA terminal is switched on. 4. Power on the safety analyzer and then press the Patient Auxiliary Current Test button on the analyzers panel to enter the patient auxiliary current test screen. 5. Repeatedly press the APPLIED PART button to measure AC and DC leakage alternatively. DC leakage reading is followed by DC. Under normal status, the patient leakage current should be no greater than 100A AC, 10A DC. Under single fault condition, the leakage current should be no greater than 500A AC, 50A DC. Contact Technical Support if the electrical safety test fails.

Accutorr V Service Manual

0070-10-0702

6 - 21

Recorder Check

Test and Calibration

6.11

Recorder Check
1. Print SpO2 Pleth waveforms. The recorder should print correctly and printout should be clear. 2. Open the recorder door and verify the monitor gives the proper message. Close recorder door before proceeding to the next step. 3. Set the recorder to print trend data. Check that the recorder prints accordingly. Contact Technical Support if the recorder test fails.

6 - 22

0070-10-0702

Accutorr V Service Manual

Test and Calibration

Software upgrade

6.12

Software upgrade
CAUTION: Disconnect the monitor from the patient and make sure that important data is saved before upgrading the monitor. Do not shut down or power off the equipment when upgrading the bootstrap program. Otherwise, it may cause the equipment to break down. Program upgrades should be performed by qualified service personnel only. After upgrading the boot program, re-upgrade the system program and other programs to ensure compatibility. Make sure the version of the upgrade package is the correct one. To obtain the latest upgrade package, contact Service. CAUTION:

CAUTION:

NOTE:

NOTE:

The following is a list of upgrade programs: Bootstrap program System program Bar code recognition configuration Multilingual library General configurations (including passwords, company logo) System functional configuration FPGA program Parameter module programs: SpO2 module (DPM), NIBP module and optional Temperature module. 1. In the MAINTENANCE dialog (see FIGURE 6-19), check the installed software version and compare it with the currently released version to determine if an upgrade is needed.

FIGURE 6-19 MAINTENANCE dialog 2. Download the upgrade software through a network to a portable PC or desktop PC. Connect the PC to the monitor using a cable plugged into the network connectors. 3. Run the upgrade software.

Accutorr V Service Manual

0070-10-0702

6 - 23

0070-10-0702 / 0070-00-0702

Rev E

November 7, 2012

You might also like