Location via proxy:   [ UP ]  
[Report a bug]   [Manage cookies]                

Lut400 Quick Start Guide Multi-Language

Download as pdf or txt
Download as pdf or txt
You are on page 1of 422

Ultrasonic Controllers

SITRANS LUT400 Compact Operating Instructions 08/2012

SITRANS

SITRANS LUT400 (HART) Quick Start Manual


This manual outlines the essential features and functions of the SITRANS LUT400 (HART 1)). We strongly advise you to acquire the detailed version of the manual so you can use your device to its fullest potential. The complete SITRANS LUT400 (HART) Operating Instructions, as well as the Communications for SITRANS LUT400 (HART) Manual, can be downloaded from the product page of our web site at: www.siemens.com/sitransLUT400. The printed manual is available from your local Siemens representative. Questions about the contents of this manual can be directed to: Siemens AG Siemens Milltronics Process Instruments 1954 Technology Drive, P .O. Box 4225 Peterborough, Ontario, Canada, K9J 7B1 Email: techpubs.smpi@siemens.com

English

Copyright Siemens AG 2012. All Rights Reserved


We encourage users to purchase authorized bound manuals, or to view electronic versions as designed and authored by Siemens Milltronics Process Instruments. Siemens Milltronics Process Instruments will not be responsible for the contents of partial or whole reproductions of either bound or electronic versions.

Disclaimer of Liability
While we have verified the contents of this manual for agreement with the instrumentation described, variations remain possible. Thus we cannot guarantee full agreement. The contents of this manual are regularly reviewed and corrections are included in subsequent editions. We welcome all suggestions for improvement. Technical data subject to change.

MILLTRONICS is a registered trademark of Siemens Milltronics Process Instruments.

Technical Support
Support is available 24 hours a day. To find your local Siemens Automation Office address, phone number, and fax number, go to: www.siemens.com/automation/partner: Click on the tab Contact to find your product group (for example: Automation Technology). Click again on the product group drop-down box to navigate to the product family. For Level: +Automation Technology > +Sensor Systems >+Process Instrumentation > +Level Measurement > +Continuous. Select Ultrasonic. For Flow: +Automation Technology > +Sensor Systems >+Process Instrumentation > +Flow Measurement. Select Ultrasonic. Select the country followed by the City/Region. Select Technical Support under Service. For on-line technical support go to: www.siemens.com/automation/support-request Enter the device name (SITRANS LUT400) or order number, then click on Search, and select the appropriate product type. Click on Next. Enter a keyword describing your issue. Then either browse the relevant documentation, or click on Next to email a description of your issue to Siemens Technical Support staff. Siemens IA/DT Technical Support Center: phone +49 (0)911 895 7222
1)

HARTis a registered trademark of HART Communication Foundation.

7ML19985XU81

SITRANS LUT400 (HART)

Page EN-1

Safety Guidelines
English
Warning notices must be observed to ensure personal safety as well as that of others, and to protect the product and the connected equipment. These warning notices are accompanied by a clarification of the level of caution to be observed. WARNING symbol relates to a caution symbol on the product, and means that failure to observe the necessary precautions can result in death, serious injury, and/or considerable material damage. WARNING symbol, used when there is no corresponding caution symbol on the product, means that failure to observe the necessary precautions can result in death, serious injury, and/or considerable material damage. Note: means important information about the product or that part of the operating manual.

FCC Conformity
US Installations only: Federal Communications Commission (FCC) rules
WARNING: Changes or modifications not expressly approved by Siemens Milltronics could void the users authority to operate the equipment. Notes: This equipment has been tested and found to comply with the limits for a Class A digital device, pursuant to Part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference when the equipment is operated in a commercial environment. This equipment generates, uses, and can radiate radio frequency energy and, if not installed and used in accordance with the instruction manual, may cause harmful interference to radio communications. Operation of this equipment in a residential area is likely to cause harmful interference to radio communications, in which case the user will be required to correct the interference at his own expense.

Page EN-2

SITRANS LUT400 (HART)

7ML19985XU81

SITRANS LUT400
WARNING: SITRANS LUT400 is to be used only in the manner outlined in this manual, otherwise protection provided by the equipment may be impaired. Note: This product is intended for use in industrial areas. Operation of this equipment in a residential area may cause interference to several frequency based communications. The Siemens SITRANS LUT400 series controllers are compact, single point, long-range ultrasonic controllers for continuous level measurement of liquids, slurries, and solids, and high accuracy monitoring of open channel flow. The series is compatible with Siemens full line of EchoMax Transducers, allowing an operating range of 0.3 to 60 meters (dependent on transducer). SITRANS LUT400 has been coupled with a backlit Local User Interface (LUI) display featuring menu driven programming. It boasts an intuitive web-based interface, and a host of wizards for plug and play performance. SITRANS LUT400 supports HART communication protocol, and several software packages for remote communications [SIMATIC PDM, AMS Device Manager, Field Communicator 375/475, and Field Device Tools (FDTs)]. It also features our next generation of Sonic Intelligence, further strengthening our industry leading measurement performance while improving ease of use.

English

Specifications
Power
AC Version

For a complete listing, see the SITRANS LUT400 (HART) Operating Instructions.

100-230 V AC Fuse: 5 x 20 mm, Slow Blow, 0.25A, 250V


Power consumption is listed at maximum.

DC Version
10-32 V DC, 10W 1 Fuse: 5 x 20 mm, Slow Blow, 1.6A, 125V

15%, 50 / 60 Hz, 36 VA (10W)1)

1)

Environmental Conditions
Altitude
2000 m max.

Ambient Temperature
-20 to +50 C (-4 to +122 F)

Installation category
II

Pollution degree
4

Relative humidity
Suitable for outdoors (only with IP65 / Type 4X / NEMA 4X Enclosure)

Approvals
Notes: The device label lists the approvals that apply to your device. Use appropriate conduit seals to maintain IP or NEMA rating. General Hazardous Non-incendive (Canada) Dust Ignition Proof (International/Europe) CSAUS/C, CE, FM, UL listed, C-TICK CSA Class I, Div. 2, Groups A, B, C, D; Class II, Div 2, Groups F, G; Class III ATEX II 3D Ex tc IIIC T70 C IP65 IECEx SIR 12.0057X SIRA 12ATEX9143X SITRANS LUT400 (HART) Page EN-3

7ML19985XU81

Installation
English
Notes: Installation must only be performed by qualified personnel, and in accordance with local governing regulations. This product is susceptible to electrostatic shock. Follow proper grounding procedures. All field wiring must have insulation suitable for at least 250V. Hazardous voltages present on transducer terminals during operation. DC input terminals shall be supplied from a source providing electrical isolation between the input and output, in order to meet applicable safety requirements of IEC 61010-1. Relay contact terminals are for use with equipment that has no accessible live parts and wiring that has insulation suitable for at least 250 V. The maximum allowable working voltage between adjacent relay contacts shall be 250 V. The non-metallic enclosure does not provide grounding between conduit connections. Use grounding type bushings and jumpers.

Mounting instructions
Mounting instructions differ for wall, pipe, DIN-rail, and remote display panel mount devices. Please follow the specific instructions for your device.

Note: When routing cable through a conduit, please follow the Cable Routing
instructions on page 9 before mounting the SITRANS LUT400.

Wall or panel mount


All configurations of the SITRANS LUT400 are shipped with a mounting back plate. The panel mount model comes with both a LUI display and a blank lid.
mounting back plate

enclosure with optional display lid clips (2 places) clips (2 places)

Note: Wall mount fasteners not included.

Page EN-4

SITRANS LUT400 (HART)

7ML19985XU81

mounting back plate

English

enclosure with blank lid clips (2 places) clips (2 places)

Note: Wall mount fasteners not included. back of enclosure

slotted features

Wall mounting the enclosure


1) 2) 3) Mark and drill four holes in the mounting surface for the four screws (customer supplied). Fasten with a screwdriver. Line up slotted features on back of device with clips on mounting back plate. Press the LUT400 flush against the back plate and slide downward to fasten in place.

mounting screw holes on back plate

Please note:
Recommended fastener size: M8 or 5/16" screw with washer of maximum 17 mm or 5/8" outside diameter Recommended mounting: mount directly to wall. If alternate mounting surface is used, it MUST be able to support four times the weight of the device.

7ML19985XU81

SITRANS LUT400 (HART)

Page EN-5

Remote mounted lid

English

remote display lid

gasket

4 panel mount fasteners provided extension cable SITRANS LUT400

Mounting the remote lid


Note: Remote mounted lid can be mounted up to 5 m from the device using two optional cable extensions (each 2.5 m in length). See full Operating Instructions. 1) Using the template provided, cut out the necessary hole for the remote LUI display lid. Place the gasket inside the lid, aligning the mounting holes. Align the back of the remote display lid with the panel hole cut-out. Mark and drill four holes in the mounting surface for the four screws (provided). Fasten with a screwdriver and wrench. Note: Recommended torque on fastening screws for good seal: 1.1 N m 10 in-lbs

2)

Please note:
Recommended mounting: mount to panel, up to 5 m from the device. If alternate mounting surface is used, it MUST be able to support four times the weight of the device. Note: Fasteners included: M5 screw, seal washer, M5 flat washer and nut. These fasteners are required to maintain IP65 rating on remote lid.

Page EN-6

SITRANS LUT400 (HART)

7ML19985XU81

Pipe mount

English

mounting back plate saddle clamp enclosure with optional display lid

u-bolt

pipe Note: Pipe mount fasteners not included.

Mounting the enclosure


1) 2) 3) Fasten the mounting back plate to the pipe using u-bolts, saddle clamps, (customer supplied) suitable to pipe diameter. Fasten bolts with a wrench. Do not over-tighten so that plate becomes twisted or bent. This may hinder ability to clip the LUT400 to the back plate. Fasten device to mounting back plate (as described in step 3 of Wall mounting the enclosure on page 5).

Please note:
Recommended mounting: directly to horizontal or vertical pipe. If alternate mounting surface is used, it MUST be able to support four times the weight of the device. Recommended pipe dimensions: maximum: 3" pipe, minimum: 3/4" pipe Recommended fastener sizes: U-Bolts: maximum: 3" pipe size with M8 or 3/8" thread minimum: 3/4" pipe size with M6 or 1/4" thread Hex Nuts: M6 or 1/4" to M8 or 3/8" Washer: maximum: 16 mm or 13/16" outside diameter.

7ML19985XU81

SITRANS LUT400 (HART)

Page EN-7

DIN-rail mount
back of enclosure

English

DIN-rail

clips (2)

slot

DIN-rail slide

Mounting the enclosure


1) 2) 3) Angle top of enclosure toward DIN-rail, and position slightly above top of rail. Move enclosure downward against DIN-rail to hook clips on back of enclosure to top of DIN-rail. Press device flush against DIN-rail to engage DIN-rail slide, which will fasten enclosure securely to DIN-rail.

Please note:
Recommended mounting: directly to horizontal DIN-rail. Required DIN-rail dimensions: TH 35-7.5 or TH 35-15 per standard IEC 60715. The DIN-rail MUST be able to support four times the weight of the SITRANS LUT400.

Removing the enclosure


1) 2) From the front of the device, place screwdriver in slot at bottom of DIN-rail slide and pry downward. This will unclip slide from bottom of DIN-rail. While holding slide down, push upward on enclosure to release clips from top of DIN-rail.

Page EN-8

SITRANS LUT400 (HART)

7ML19985XU81

Preparation for cable entry


Cables can be routed through conduit or enter the enclosure through cable glands. Follow steps 1 to 5 below to first uncover cable entry holes, then complete steps for use with conduit, or with cable glands. 1) 2) 3) 4) 5) Ensure enclosure lid is closed and fastener screws are locked. Place tip of screwdriver into groove on the outer diameter of the knock-out tab (see illustration that follows). Hit the end of the screwdriver with palm of hand to knock out entry hole. Loosen screws and remove enclosure lid. Remove plastic piece(s) that covered entry holes from enclosure. Be careful not to damage the electronics with static electricity, or the tools used to knock out entry holes.
SITRANS LUT400

English

cable entry holes

knock-out tab (3 places)

groove for screwdriver (3 places)

Notes: When cable entry hole knockout tabs have been removed, the entry hole is 21.4 mm to 21.6 mm in diameter. M20 cable glands (20 mm in diameter), and 1/2" NPT conduit (21.3 mm in diameter) fit this entry hole. Caution should be taken when selecting appropriate seal for entry holes. Flat gasket is recommended (instead of O-ring). If alternate cable glands are used, it is the customers responsibility to maintain IP65 rating of entry holes.

Cable routed through conduit


(continued from steps 1 to 5 above) 6) After preparing for cable entry in steps 1 to 5 above, attach the conduit to the enclosure using only suitable size fittings approved for water-tight applications. (Conduit size is 1/2" NPT.) 7) Replace enclosure lid and tighten screws.

Cable exposed and entering through the cable glands


(continued from steps 1 to 5 above) 6) After preparing for cable entry in steps 1 to 5 above, unscrew the glands and attach them firmly to the enclosure. 7) Thread the cables through the glands. Ensure the power cable is kept separated from the signal cables and then wire the cables to the terminal blocks. 8) Tighten the glands to form a good seal. 9) Replace enclosure lid and tighten screws.

7ML19985XU81

SITRANS LUT400 (HART)

Page EN-9

Connecting
Safety notes for connection
For complete wiring instructions, please refer to the full Operating Instructions. Verify that all system components are installed in accordance with instructions. Connect all cable shields to the LUT400 shield terminals (denoted on device with symbol ). To avoid differential ground potentials ensure cable shields are properly connected to ground. Keep exposed conductors on shielded cables as short as possible to reduce noise on the line caused by stray transmissions and noise pickup.

English

Connecting SITRANS LUT400


WARNINGS:

Check the device label on your instrument, to verify the approval rating. Use appropriate conduit seals to maintain IP or NEMA rating.
Notes: Separate cables and conduits may be required to conform to standard instrumentation wiring practices or electrical codes. To access the wiring compartment: 1) 2) 3) Loosen 1/4 turn locking screws. Lift lid up and to the left on its hinges. The lid can remain open connected by hinges, or it can be unclipped from the hinges and set to one side, to access wiring compartment.

4) 5) 6)

Make all connections as per instructions that follow. When wiring complete, replace device lid. Tighten locking screws.

Page EN-10

SITRANS LUT400 (HART)

7ML19985XU81

Wiring compartment
The terminal board on the LUT400 allows all inputs and outputs to be connected simultaneously. Terminal strips can be removed to improve ease of wiring.

English

CAUTION: Ensure the terminal strips are terminated to the correct location during re-installation. Failure to do so may result in damage to the device, or external equipment that is attached. Note: Recommended torque on terminal clamping screws. 0.56 - 0.79 N m 5 - 7 in-lbs Please do not overtighten the screws. ower
TS-3 TRANSDUCER DISCRETE INPUTS INPUT POWER

power supply
14

8 1

9 2

10

11 0.25 A, 250 V SLOW BLOW

12 L1

13 L2/N 100 - 230 V

DC SYNC

SHLD

WHT

BLK

BLK

WHT

terminal blocks
mA OUT/HART BATTERY BR2032

battery

RELAYS

DISPLAY 15 16 RLY 1 17 18 19 20 21 22 23 24

RLY 2

RLY 3 ACT

PAS SERVICE

USB connection

display cable

mA HART connection

Power
WARNINGS: The DC input terminals shall be supplied from a source providing electrical isolation between the input and output, in order to meet the applicable safety requirements of IEC 61010-1. All field wiring must have insulation suitable for rated voltages. Important! Before applying power to the SITRANS LUT400 for the first time, ensure any connected alarm/control equipment is disabled until satisfactory system operation and performance is verified.

7ML19985XU81

SITRANS LUT400 (HART)

Page EN-11

The SITRANS LUT400 is available in AC or DC power models.

AC Power English
INPUT POWER

DC Power
INPUT POWER

12

13

14

12 10 - 32 V

13

14

L1 L2/N 100 - 230 V

AC: 100-230 V AC 15%, 50/60 Hz, 36 VA (10W)

DC: 10-32 V DC, 10W

Note: Make sure device is connected to a reliable ground. 1) 2) 3) To wire for power, strip the cable jacket for approximately 70 mm (2.75") from the end of the cable, and thread the wires through the gland 1). Connect the wires to the terminals as shown: the polarity is identified below the terminal block. Ground the device according to local regulations. Notes for AC power connections: The equipment must be protected by a 15 A fuse, or circuit breaker in the building installation. A circuit breaker or switch in the building installation, marked as the disconnect switch, must be in close proximity to the equipment and within easy reach of the operator.

The Battery
The SITRANS LUT400 is supplied with one battery installed. (See illustration on page 11 for battery location.) The battery (BR2032) has a life expectancy of ten years, and is affected by ambient temperature. If the LUT400 loses input power, the battery will maintain operation of the devices real time clock until power has been restored. (See full Operating Instructions for battery replacement.)

1)If cable is routed through conduit, use only approved suitable-size hubs for waterproof applications.

Page EN-12

SITRANS LUT400 (HART)

7ML19985XU81

Transducers
Warning: Hazardous voltage present on transducer terminals during operation. Run the transducer cable in a grounded metal conduit, separate from other wiring (except TS-3 temperature sensor wiring, if applicable). Notes: Do not use coaxial cable with SITRANS LUT400 due to high voltage transmitted on shield of coaxial cable Do not connect the LUT400 shield and white transducer wires together; wire to separate terminals Disregard older transducer manuals that recommend these practices
TS-3 TRANSDUCER

English

SHLD

BLK

WHT

Relays
Relay contacts are shown in the de-energized position. All relays can be configured as positive or negative logic (see parameter 2.8.11.Relay Logic in full Operating Instructions).
RELAYS

BLK

WHT

Power Failure
Relays 2, 3 are normally open. Relay 1 can be wired either normally open or normally closed. In the event of loss of input power, relays will revert to their normal states.

15

16 RLY 1

17

18

19

20

21

RLY 2

RLY 3

Relay Ratings
one Form C (NO or NC) relay (relay 1), 1A at 250 V AC, noninductive, 3A at 30 V DC two Form A (NO) relays (relays 2,3), 5A at 250 V AC, non-inductive, 3A at 30 V DC

7ML19985XU81

SITRANS LUT400 (HART)

Page EN-13

Connecting HART mA Output (HART) English


mA OUT/HART

For ACTIVE HART connection (using LUT400 integral power supply), connect terminals 22 and 23.

22

23

24

For PASSIVE HART connection (using external power supply), connect terminals 23 and 24.

ACT

PAS

For more information, consult the mA output parameters (2.5.Current Output) in the parameter reference section of the full Operating Instructions.

Connecting in hazardous area installations


See page 35.

Commissioning SITRANS LUT400


We recommend that you configure your application in the following order: First, run the appropriate Quick Start Wizard for your application (Level, Volume, Flow). Next, setup pumps via the Pump Control Wizard (if applicable). Lastly, configure alarms, or other controls, totalizers and samplers, referencing the respective parameters (see full Operating Instructions). There are two ways to access the Quick Start Wizards: locally (see Quick Start Wizards via LUI on page 17) from a remote location (see Other Quick Start Wizards: on page 17)

Activating SITRANS LUT400


Power up the device. SITRANS LUT400 automatically starts up in Measurement mode. A transition screen showing first the Siemens logo and then the current firmware revision of the LUI is displayed while the first measurement is being processed. The first time the device is configured, you will be prompted to select a language (English, German, French, Spanish or Chinese). The correct date and time should be set prior to configuring the device. (To modify, set parameter 2.14.Date and Time. See full Operating Instructions for details.)

Page EN-14

SITRANS LUT400 (HART)

7ML19985XU81

The LCD Display Measurement mode display: Normal operation 1) Level


1 10
SITRANS LUT400 LEVEL

2
1/2

4 5

English

18.91
1

6
2

5 SAVING PARAMETERS

Flow
1 10 11 12
SITRANS LUT400 FLOW HEAD
1 2 3

3
1/2
L/S M

4 5 6

500313..
60.000

DAILY TOTAL,...

999999.99

5 SAVING PARAMETERS

1 tag 2 measured value (level, space, distance, volume, flow, or head) 3 value being displayed [Primary Variable (PV)=1 of 2, Secondary Variable (SV)=2 of 2] 4 toggle indicator1) for PV or SV 5 units 6 bar graph indicates level 7 secondary region indicates configured relays (left) and discrete inputs (right) 8 text area displays status messages 9 device status indicator 10 selected (primary) sensor mode: level, space, distance, volume, head, or flow 11 secondary sensor mode = head when primary sensor mode = flow 12 totalizer values: display alternates between daily totalizer and running totalizer

Fault present
0 LOE

8 text area displays a fault code and an error message 9 service required icon appears

Program mode display Navigation view


A visible menu bar indicates the menu current TRANSDUCER 2.1.6 menu/ list is too long to display all items. NO TRANSDUCER parameter parameter XRS-5 name number The depth of the item band on the XPS-10 item menu bar indicates the length of the XPS-15 current XCT-8 band menu list: a deeper band indicates item XCT-12 XPS-30 menu fewer items. bar The position of the item band indicates the approximate position of the current item in the list. A band halfway down the menu bar indicates the current item is halfway down the list. A deeper band indicates fewer items.

1)Press

or

to switch.

7ML19985XU81

SITRANS LUT400 (HART)

Page EN-15

Parameter view English


menu in which parameter is located parameter name
SENSOR
UNITS SENSOR MODE SENSOR MODE SECONDARY

Edit view
2.1.6

parameter number parameter value/ selection

TRANSDUCER
NO TRANSDUCER XPS-10 XPS-15 XCT-8 XCT-12 XPS-30

2.1.6

XRS-5

TRANSDUCER

FREQUENCY

XRS-5

PROGRAM mode
Using the device local push buttons, press Scroll through the menu using , , , to enter Program mode and open menu level 1. .

Selecting a listed option


a) Navigate to the desired parameter. b) Press RIGHT arrow to open Edit mode. The current selection is highlighted.

c) Scroll to a new selection. d) Press RIGHT arrow the new selection. to accept it. The LCD returns to parameter view and displays

Changing a numeric value


a) Navigate to the desired parameter, for example, Low Calibration Point (2.2.1.). b) When selected, the current value is displayed. c) Press RIGHT arrow to open Edit mode. The cursor position is highlighted. d) Use LEFT and RIGHT arrow digit position you wish to change. to move cursor to and DOWN arrow to
LOW CALIB. PT. 2.2.1

ESC

+ _ _ _ _ _ _ 2. 0 M

OK

e) As each digit is highlighted (selected), use the UP increase or decrease the digit respectively. f) While decimal point is selected, use UP position.

and DOWN arrow

to shift decimal continually until ESC

g) To escape without saving your changes, press LEFT arrow is highlighted. Press LEFT arrow

again to escape without saving changes.

Otherwise, when new parameter value is correct, press RIGHT arrow continually until OK is highlighted. h) Press RIGHT arrow to accept the new value. The LCD returns to parameter view and displays the new selection. Review for accuracy.

Page EN-16

SITRANS LUT400 (HART)

7ML19985XU81

Quick Start Wizards


To configure the SITRANS LUT400 for applications of level, volume (standard vessel shapes), or flow, use the Quick Start Wizards via LUI described below. A wizard for applications employing more complex vessel shapes is available via SIMATIC PDM. See Quick Start Wizards via SIMATIC PDM on page 33.

English

Other Quick Start Wizards:


Other Quick Start Wizards using various software packages are also available 1): SIMATIC PDM (HART) AMS (HART) FC375/475 (HART) FDTs (HART) Parameter Configuration Charts (to record data prior to configuring device via Quick Start Wizard) are available on our website. Go to www.siemens.com/sitransLUT400 > Support > Application Guides.

Quick Start Wizards via LUI


1) 2) Press to enter Program mode. Choose Wizards (1.), Quick Start (1.1.), and then the appropriate quick start: QS Level (1.1.1.), QS Volume (1.1.2.), or QS Flow (1.1.3.). [The QS Flow wizard will display on LUI for LUT430 (Pump and Flow), and LUT440 (OCM) configured models only.] Follow the steps then choose Finish to save Quick Start parameter changes and return to Program menu, then press three times to return to Measurement mode.

3)

Notes: The Quick Start Wizard settings are inter-related and changes apply only after you choose Finish in the final step. Do not use the Quick Start Wizard to modify individual parameters: see instead Parameter section of full Operating Instructions. (Perform customization for your application only after the Quick Start has been completed.). For more information on any wizard parameter see full Operating Instructions.

1. Wizards 1.1. Quick Start 1.1.1. QS Level


Use this wizard to configure simple level applications. a) Press RIGHT arrow MENU. b) Press RIGHT arrow c) Press RIGHT arrow to activate PROGRAM mode and open menu level 1: MAIN two times to navigate to menu item 1.1.1. to open QS Level.

1)See Communications for SITRANS LUT400 (HART) Manual

7ML19985XU81

SITRANS LUT400 (HART)

Page EN-17

d) At each step, press DOWN arrow to accept default values and move directly to the next item,

QUICK START LEVEL


TRANSDUCER XRS-5
BACK CANCEL NEXT EDIT

English

or RIGHT arrow to open Edit mode: the current selection is highlighted. e) Scroll to desired item and press RIGHT arrow to store the change, then press DOWN arrow f) to continue.

QUICK START LEVEL


TRANSDUCER

NO TRANSDUCER XPS-10 XPS-15 XCT-8 XCT-12

XRS-5

At any time, you can press UP arrow wizard. Start of QS Level Wizard

to go back, or LEFT arrow

to cancel the

Shows the type of Wizard to be executed.

Options
Transducer

CANCEL, START

Specifies the Siemens transducer connected to the device.

Options
Operation

NO TRANSDUCER, XRS-5, XPS-10, XPS-15, XCT-8, XCT-12, XPS-30, XPS-40, XLT-30, XLT-60, STH Default: NO TRANSDUCER

Sets the type of measurement (and the corresponding mA output) required for the application.
sensor reference point High Calibration Point distance space

level Low Calibration Point

Far Range

Mode
LEVEL SPACE DISTANCE

Description
* Height of material Distance to material surface

Reference point
Low Calibration Point (process empty level) High Calibration Point (process full level) Sensor Reference Point (continued on next page)

Page EN-18

SITRANS LUT400 (HART)

7ML19985XU81

(continued from previous page)

Mode

Description
Do NOT select. If Operation value displays as OTHER, the device is configured as a level controller, but has been set previously to a mode other than LEVEL, SPACE, or DISTANCE. The operation mode must be set to LEVEL, SPACE, or DISTANCE to proceed with the QS Level Wizard.

Reference point

English

OTHER

Temperature Source

Source of the temperature reading used to adjust the speed of sound.

Options

TRANSDUCER, FIXED TEMPERATURE, EXTERNAL TS-3, AVERAGE OF SENSORS Default: TRANSDUCER

Fixed Temperature

Use this feature if a temperature sensing device is not used.

Value

Range: -100.0 to +150.0 C Default: +20.0 C

This parameter is only displayed if FIXED TEMPERATURE selected for Temperature Source. Units

Sensor measurement units.

Options

M, CM, MM, FT, IN Default: M

Note: For the purpose of this example, all values are assumed to be in meters (m).
High Calibration Point

Distance from Sensor Reference Point to High Calibration Point: usually process full level.

Value

Range: 0.000 to 60.000 Default: 0.000

Low Calibration Point

Distance from Sensor Reference Point to Low Calibration Point: usually process empty level.

Value

Range: 0.000 to 60.000 Default: 60.000

7ML19985XU81

SITRANS LUT400 (HART)

Page EN-19

Response Rate

Sets the reaction speed of the device to measurement changes in the target range.

Notes: English

Response Rate can only be set through the Quick Start Wizard, and any changes made to Fill Rate per Minute (2.3.1.), Empty Rate per Minute (2.3.2.), or Damping Filter (2.3.3.) parameters following the completion of the wizard will supersede the Response Rate setting. Response Rate always displays in m/minute.
SLOW (0.1 M/MIN) MEDIUM (1.0 M/MIN) FAST (10 M/MIN) Default: SLOW (0.1 M/MIN)

Options

Use a setting just faster than the maximum filling or emptying rate (whichever is greater). Slower settings provide higher accuracy, faster settings allow for more rapid level fluctuations. End of QS Level Wizard

For QS to be successful, all changes must be applied.

Options

BACK, CANCEL, FINISH (Display returns to 1.1 Quick Start menu when Quick Start is successfully completed or cancelled. If CANCEL is selected, no changes are written to the device.)

To transfer Quick Start values to the device and return to Program menu, press DOWN arrow (Finish). Then press LEFT arrow Measurement mode. three times to return to

1.1.2. QS Volume
Use this wizard to configure volume applications employing standard vessel shapes. a) Press RIGHT arrow MENU. b) Press RIGHT arrow c) Press DOWN arrow to activate PROGRAM mode and open menu level 1: MAIN two times to navigate to menu item 1.1.1. and RIGHT arrow to open QS Volume.
QUICK START VOLUME
TRANSDUCER XRS-5
BACK CANCEL NEXT EDIT

d) At each step, press DOWN arrow to accept default values and move directly to the next item, or RIGHT arrow to open Edit mode: the current selection is highlighted. e) Scroll to desired item and press RIGHT arrow store the change, then press DOWN arrow continue. f) At any time, you can press UP arrow LEFT arrow Page EN-20 to cancel the wizard. SITRANS LUT400 (HART) to to

QUICK START VOLUME


TRANSDUCER

NO TRANSDUCER XPS-10 XPS-15 XCT-8 XCT-12

XRS-5

to go back, or

7ML19985XU81

Start of QS Volume Wizard

Shows the type of Wizard to be executed.

Options
Transducer

CANCEL, START

English

Specifies the Siemens transducer connected to the device.

Options

NO TRANSDUCER, XRS-5, XPS-10, XPS-15, XCT-8, XCT-12, XPS-30, XPS-40, XLT-30, XLT-60, STH Default: NO TRANSDUCER

Temperature Source

Source of the temperature reading used to adjust the speed of sound.

Options

TRANSDUCER, FIXED TEMPERATURE, EXTERNAL TS-3, AVGERAGE OF SENSORS Default: TRANSDUCER

Fixed Temperature

Use this feature if a temperature sensing device is not used.

Value

Range: -100.0 to +150.0 C Default: +20.0 C

This parameter is only displayed if FIXED TEMPERATURE selected for Temperature Source. Vessel Shape

Defines the vessel shape and allows the SITRANS LUT400 to calculate volume instead of level. If NONE is selected, no volume conversion is performed. Select the vessel shape matching the monitored vessel or reservoir.

Options

NONE, LINEAR, CYLINDER, PARABOLIC BOTTOM, HALF SPHERE BOTTOM, FLAT SLOPED BOTTOM, PARABOLIC ENDS, SPHERE, CONICAL BOTTOM, CURVE TABLE, LINEAR TABLE Default: LINEAR

If CURVE TABLE or LINEAR TABLE selected, enter values for level and volume breakpoints after completing the wizard. (See parameter 2.6.7.Table 1-8 in full Operating Instructions.) Units

Sensor measurement units.

Options

M, CM, MM, FT, IN Default: M

Note: For the purpose of this example, all values are assumed to be in meters (m).

7ML19985XU81

SITRANS LUT400 (HART)

Page EN-21

High Calibration Point

Distance from Sensor Reference Point to High Calibration Point: usually process full level.

English

Value

Range: 0.000 to 60.000 Default: 0.000

Low Calibration Point

Distance from Sensor Reference Point to Low Calibration Point: usually process empty level.

Value
Response Rate

Range: 0.000 to 60.000 Default: 60.000

Sets the reaction speed of the device to measurement changes in the target range.

Notes:

Response Rate can only be set through the Quick Start Wizard, and any changes made to Fill Rate per Minute (2.3.1.) or Empty Rate per Minute (2.3.2.) parameters following the completion of the wizard will supersede the Response Rate setting. Response Rate always displays in m/minute.
SLOW (0.1 M/MIN) MEDIUM (1.0 M/MIN) FAST (10 M/MIN) Default: SLOW (0.1 M/MIN)

Options

Use a setting just faster than the maximum filling or emptying rate (whichever is greater). Slower settings provide higher accuracy, faster settings allow for more level fluctuations. Dimension A

The height of the vessel bottom when the bottom is conical, pyramidal, parabolic, spherical, or flat -sloped..

Value
Dimension L

Range: 0.000 to 99.999 Default: 0.000

Length of the cylindrical section of a horizontal parabolic end vessel.

Value
Volume Units

Range: 0.000 to 99.999 Default: 0.000

Determines volume measurement units.

Options

L, USGAL, IMPGAL, CUM, USER DEFINED * Default: L

* If USER DEFINED option selected, the value must be set after completing the wizard. See parameter 2.6.6.User Defined Unit in full Operating Instructions. Page EN-22 SITRANS LUT400 (HART) 7ML19985XU81

Maximum Volume

The maximum volume of the vessel. Enter the vessel volume corresponding to the High Calibration Point. For example, if your maximum vessel volume is 8000 L, enter a value of 8000.

English

Value

Range: 0.0 to 9999999 Default: 100.0

End of QS Volume Wizard

For QS to be successful, all changes must be applied.

Options

BACK, CANCEL, FINISH (Display returns to 1.1 Quick Start menu when Quick Start is successfully completed or cancelled. If CANCEL is selected, no changes are written to the device.)

To transfer Quick Start values to the device and return to Program menu, press DOWN arrow (Finish). Then press LEFT arrow Measurement mode. three times to return to

1.1.3. QS Flow
Use this wizard to configure simple flow applications. [Visible on LUT430 (Pump and Flow), and LUT440 (OCM) configured models only.] a) Press RIGHT arrow MENU. b) Press RIGHT arrow c) Press DOWN arrow to activate PROGRAM mode and open menu level 1: MAIN two times to navigate to menu item 1.1.1. two times and RIGHT arrow to open QS Flow.
QUICK START FLOW
TRANSDUCER XRS-5
BACK CANCEL NEXT EDIT

d) At each step, press DOWN arrow to accept default values and move directly to the next item, or RIGHT arrow to open Edit mode: the current selection is highlighted. e) Scroll to desired item and press RIGHT arrow to store the change, then press DOWN arrow f) to continue. to

QUICK START FLOW


TRANSDUCER

NO TRANSDUCER XPS-10 XPS-15 XCT-8 XCT-12

XRS-5

At any time, you can press UP arrow go back, or LEFT arrow Start of QS Flow Wizard

to cancel the wizard.

Shows the type of Wizard to be executed.

Options

CANCEL, START

7ML19985XU81

SITRANS LUT400 (HART)

Page EN-23

Transducer

Specifies the Siemens transducer connected to the device.

English

Options

NO TRANSDUCER, XRS-5, XPS-10, XPS-15, XCT-8, XCT-12, XPS-30, XPS-40, XLT-30, XLT-60, STH Default: NO TRANSDUCER

Temperature Source

Source of the temperature reading used to adjust the speed of sound.

Options

TRANSDUCER, FIXED TEMPERATURE, EXTERNAL TS-3, AVGERAGE OF SENSORS Default: TRANSDUCER

Fixed Temperature

Source of the temperature reading used to adjust the speed of sound.

Value

Range: -100.0 to +150.0 C Default: +20.0 C

This parameter is only displayed if FIXED TEMPERATURE selected for Temperature Source. Primary Measuring Device

Defines the primary measuring device (PMD) to be used in the application.


EXPONENTIAL DEVICES, RECTANGULAR FLUME BS-3680, ROUND NOSE HORIZONTAL CR. BS-3680, TRAPEZOIDAL FLUME BS-3680, U-FLUME BS3680, FINITE CREST WEIR BS-3680, THIN PLATE RECT. WEIR BS-3680, THIN PLATE V-NOTCH WEIR BS-3680, RECT. WEIR CONTRACTED, ROUND PIPE, PALMER BOWLUS FLUME, H-FLUME, OTHER* Default: EXPONENTIAL DEVICES

Options

* Option will be set to OTHER if the wizard was run previously via HART software tool (such as SIMATIC PDM), and the device was set to OFF or UNIVERSAL HEAD VS. FLOW. If this is initial configuration, the PMD can only be set for no calculation (OFF), or for linearization (UNIVERSAL HEAD VS. FLOW) via HART software tools (SIMATIC PDM, AMS, FC375/475). See full Operating Instructions. Method of Flow Calculation

Sets the method of flow calculation.

Options
Units

ABSOLUTE, RATIOMETRIC Default: ABSOLUTE

Sensor measurement units.

Options

M, CM, MM, FT, IN Default: M

Note: For the purpose of this example, all values are assumed to be in meters (m).

Page EN-24

SITRANS LUT400 (HART)

7ML19985XU81

High Calibration Point

Distance from Sensor Reference Point to High Calibration Point: usually process full level.

Value

Range: 0.000 to 60.000 Default: 0.000

English

Low Calibration Point

Distance from Sensor Reference Point to Low Calibration Point: usually process empty level.

Value
Response Rate

Range: 0.000 to 60.000 Default: 60.000

Sets the reaction speed of the device to measurement changes in the target range.

Notes:

Response Rate can only be set through the Quick Start Wizard, and any changes made to Fill Rate per Minute (2.3.1.) or Empty Rate per Minute (2.3.2.) parameters following the completion of the wizard will supersede the Response Rate setting. Response Rate always displays in m/minute.
SLOW (0.1 M/MIN) MEDIUM (1.0 M/MIN) FAST (10 M/MIN) Default: SLOW (0.1 M/MIN)

Options

Use a setting just faster than the maximum filling or emptying rate (whichever is greater). Slower settings provide higher accuracy, faster settings allow for more level fluctuations. Calculation factors:

Notes:

The following five parameters will display in the wizard based on the PMD selected above. These parameters are used in the flow calculation formula (see full Operating Instructions for details). Flow Exponent (PMD = EXPONENTIAL DEVICES)

The exponent for the flow calculation formula.

Value

Range: -999.000 to 9999.000 Default: 1.550

7ML19985XU81

SITRANS LUT400 (HART)

Page EN-25

K Factor (PMD = EXPONENTIAL DEVICES)

English

The constant used in the flow calculation formula for absolute calculation of an exponential device only.

Value
V-Notch Angle

Range: -999.000 to 9999.000 Default: 1.000

(PMD = THIN PLATE V-NOTCH WEIR)

The V-Notch angle used in the flow calculation formula.

Value
Slope

Range: 25.000 to 95.000 Default: 25.000

(PMD = TRAPEZOIDAL FLUME or ROUND PIPE)

The Flow Slope used in the flow calculation formula.

Value

Range: -999.000 to 9999.000 Default: 0.000

Roughness Coefficient (PMD = ROUND PIPE)

The Flow Roughness Coefficient used in the flow calculation formula.

Value

Range: -999.000 to 9999.000 Default: 0.000

Page EN-26

SITRANS LUT400 (HART)

7ML19985XU81

PMD Dimensions

Notes:

For each PMD excluding Exponential Devices, and Other, you must enter up to four dimensions. In the wizard, you will be prompted for each dimension required for the PMD selected, and the respective PMD dimension name will be displayed. PMD selected Wizard dimension name (parameter menu reference)
APPROACH WIDTH B (2.15.4.5. OCM Dimension 1) THROAT WIDTH B (2.15.4.6. OCM Dimension 2) HUMP HEIGHT P (2.15.4.7. OCM Dimension 3) THROAT LENGTH L (2.15.4.8. OCM Dimension 4)

English

Rectangular Flume BS-3680

Round Nose Horizontal Crest Weir BS-3680


CREST WIDTH B (2.15.4.5. OCM Dimension 1) CREST HEIGHT P (2.15.4.6. OCM Dimension 2) CREST LENGTH L (2.15.4.7. OCM Dimension 3)

Trapezoidal Flume BS-3680


APPROACH WIDTH B (2.15.4.5. OCM Dimension 1) THROAT WIDTH B (2.15.4.6. OCM Dimension 2) HUMP HEIGHT P (2.15.4.7. OCM Dimension 3) THROAT LENGTH L (2.15.4.8. OCM Dimension 4)

U-Flume BS-3680
APPROACH DIAMETER DA (2.15.4.5. OCM Dimension 1) THROAT DIAMETER D (2.15.4.6. OCM Dimension 2) HUMP HEIGHT P (2.15.4.7. OCM Dimension 3) THROAT LENGTH L (2.15.4.8. OCM Dimension 4)

Finite Crest Weir BS-3680


CREST WIDTH B (2.15.4.5. OCM Dimension 1) CREST HEIGHT P (2.15.4.6. OCM Dimension 2) CREST LENGTH L (2.15.4.7. OCM Dimension 3)

Thin Plate Rectangular Weir BS-3680


APPROACH WIDTH B (2.15.4.5. OCM Dimension 1) CREST WIDTH B (2.15.4.6. OCM Dimension 2) CREST HEIGHT P (2.15.4.7. OCM Dimension 3)

Rectangular Weir Contracted


CREST WIDTH B (2.15.4.5. OCM Dimension 1) (continued on next page)

7ML19985XU81

SITRANS LUT400 (HART)

Page EN-27

PMD selected (contd) Round Pipe

Wizard dimension name (parameter menu reference)


PIPE INSIDE DIAMETER D (2.15.4.5. OCM Dimension 1)

English

Palmer Bowlus Flume


MAXIMUM FLUME WIDTH HMAX (2.15.4.5. OCM Dimension 1)

H-Flume
MAXIMUM LISTED HEAD HMAX (2.15.4.5. OCM Dimension 1)

Maximum Head

The maximum level value associated with the PMD.

Value
Zero Head Offset

Range: 0.000 to 60.000 Default: 60.000

The difference (positive or negative) between Low Calibration Point and zero head (level at zero flow).

Value
Flowrate Units

Range: -60.000 to 60.000 Default: 0.000

The volume units used to display total flow.

Options

L/S, L/MIN, CUFT/S, CUFT/D, GAL/MIN, GAL/D, IMPGAL/MIN, IMPGAL/D, CUM/H, CUM/D, USER DEFINED * Default: L/S

* If USER DEFINED option selected, the value must be set after completing the wizard. See parameter 2.6.6.User Defined Unit in full Operating Instructions. Maximum Flow at 20 mA

The maximum flowrate.

Value

Range: -999 to 9999999 Default: 100

Flowrate Decimal

The maximum number of decimal units to be displayed.

Options
Low Flow Cutoff

NO DIGITS, 1 DIGIT, 2 DIGITS, 3 DIGITS Default: NO DIGITS

Eliminates totalizer activity for head levels at or below the cutoff value.

Value

Range: 0.000 to 60.000 Default: 0.000

Page EN-28

SITRANS LUT400 (HART)

7ML19985XU81

End of QS Flow Wizard

For QS to be successful, all changes must be applied.

Options

BACK, CANCEL, FINISH (Display returns to 1.1.1 Quick Start menu when Quick Start is successfully completed or cancelled. If CANCEL is selected, no changes are written to the device.)

English

To transfer Quick Start values to the device and return to Program menu, press DOWN arrow (Finish). Then press LEFT arrow three times to return to Measurement mode. completion of the wizard to ensure best accuracy. See full Operating Instructions for details.

Note: It is strongly recommended that an 2.15.2.Auto Zero Head be performed after

1.2. Pump Control


Use this wizard to configure pumps if they will be used in your application. Be sure to first complete the applicable Quick Start Wizard. a) Press RIGHT arrow MENU. b) Press RIGHT arrow c) Press DOWN arrow to activate PROGRAM mode and open menu level 1: MAIN to navigate to menu item 1.1. and RIGHT arrow to open Pump Control.

d) At each step, press DOWN arrow the next item, or RIGHT arrow

to accept default values and move directly to

to open Edit mode: the current selection is highlighted. to store the change, then press

e) Scroll to desired item and press RIGHT arrow DOWN arrow f) to continue.

At any time, you can press UP arrow wizard. Start of Wizard - Pump Control

to go back, or LEFT arrow

to cancel the

Shows the type of Wizard to be executed.

Options

CANCEL, START

Number of Pumps

Select the number of pumps to be used with pump control.

Options

NONE, 2 Default: NONE

If set to NONE, pump control is disabled.

7ML19985XU81

SITRANS LUT400 (HART)

Page EN-29

Relay Pump 1

Selects the relay assigned to Pump 1.

English

Options
Relay Pump 2

RELAY 2, RELAY 3 Default: RELAY 2

View only. Automatically sets the relay assigned to Pump 2 based on relay selected for Pump 1 in previous step.

Options (view only)

If Relay Pump 1 = RELAY 2, then Relay Pump 2 = RELAY 3 If Relay Pump 1 = RELAY 3, then Relay Pump 2 = RELAY 2

Pump Control Mode

Sets the control algorithm used to trip the relay.

LUT420 Level Controller:


ALTERNATE DUTY ASSIST, ALTERNATE DUTY BACKUP

LUT430 Pump and Flow Controller: Options supported per model


ALTERNATE DUTY ASSIST, ALTERNATE DUTY BACKUP , SERVICE RATIO DUTY ASSIST, SERVICE RATIO DUTY BACKUP , FIXED DUTY ASSIST, FIXED DUTY BACKUP

LUT440 High Accuracy OCM:


ALTERNATE DUTY ASSIST, ALTERNATE DUTY BACKUP , SERVICE RATIO DUTY ASSIST, SERVICE RATIO DUTY BACKUP , FIXED DUTY ASSIST, FIXED DUTY BACKUP Default (all models): ALTERNATE DUTY ASSIST

Service Ratio Pump 1

Selects pump usage based on the RUN time ratio rather than last used.

Value

Range: 0 to 255 Default: 1

This parameter displays only if a Service Ratio algorithm is selected for Pump Control Mode. Service Ratio Pump 2

Selects pump usage based on the RUN time ratio rather than last used.

Value

Range: 0 to 255 Default: 1

This parameter displays only if a Service Ratio algorithm is selected for Pump Control Mode.

Page EN-30

SITRANS LUT400 (HART)

7ML19985XU81

Run Time Relay 2

Set the amount of time that pump Relay 2 has run, defined in hours.

Value

Range: 0 to 999999 Default: 0

English

Use the default value for new pumps, or set this value for existing pumps with accumulated run time. (This parameter displays only if a Service Ratio algorithm is selected for Pump Control Mode.) Run Time Relay 3

Set the amount of time that pump Relay 3 has run, defined in hours.

Value

Range: 0 to 999999 Default: 0

Use the default value for new pumps, or set this value for existing pumps with accumulated run time. (This parameter displays only if a Service Ratio algorithm is selected for Pump Control Mode.) ON Setpoint Pump 1

The level at which Pump 1 turns ON, defined in 2.1.1.Units.

Value

Range: 0.000 to 99999.000 Default: 0.000

ON Setpoint Pump 2

The level at which Pump 2 turns ON, defined in 2.1.1.Units.

Value

Range: 0.000 to 99999.000 Default: 0.000

OFF Setpoint Pump 1

The level at which Pump 1 turns OFF, defined in 2.1.1.Units.

Value

Range: 0.000 to 99999.000 Default: 0.000

OFF Setpoint Pump 2

The level at which Pump 2 turns OFF, defined in 2.1.1.Units.

Value

Range: 0.000 to 99999.000 Default: 0.000

End of Wizard - Pump Control

For Wizard to be successful, all changes must be applied.

Options

BACK, CANCEL, FINISH (Display returns to Pump Control menu when Wizard is successfully completed or cancelled. If CANCEL is selected, no changes are written to the device.)

To transfer values to the device and return to Program menu, press DOWN arrow (Finish). Then press LEFT arrow two times to return to Measurement mode.

7ML19985XU81

SITRANS LUT400 (HART)

Page EN-31

SITRANS LUT400 Communications: HART


SITRANS LUT400 can be configured over the HART network using the HART Communicator 375/475 by Emerson (see full Operating Instructions for details), or a software package. The recommended software package is the SIMATIC Process Device Manager (PDM) by Siemens. Details are outlined in an application guide Working with HART, which can be downloaded from the product page of our website. Go to: www.siemens.com/sitransLUT400 under Support and click on Application Guides.

English

Remote operation
SITRANS LUT400 supports several software tools for operation via remote communications: PC running SIMATIC PDM PC running Emerson AMS Device Manager PC running a web browser PC running a Field Device Tool (FDT) Field Communicator 375/475 (FC375/FC475). Below you will find the basic information required to use these tools with your SITRANS LUT400. Further details for each tool are available in the Communications for SITRANS LUT400 (HART) Manual.

SIMATIC PDM
SIMATIC PDM is a software package used to commission and maintain SITRANS LUT400 and other process devices. You can find more information at www.siemens.com/simatic-pdm. Check the support page of our website to make sure you have the latest version of SIMATIC PDM, the most recent Service Pack (SP) and the most recent hot fix (HF). Go to: http://support.automation.siemens.com/WW/ llisapi.dll?func=cslib.csinfo&lang=en&objiD=10806857&subtype=133100

Update Electronic Device Description (EDD)


You can locate the EDD in Device Catalog, under Sensors/Level/Echo/Siemens AG/SITRANS LUT400. (The EDD is written for forward compatibility.) As a guideline to locate the correct EDD, the major and minor numbers should match between the EDD revision and the Firmware revision in the device (e.g. major and minor numbers in bold text: 1.00.00-02). The EDD revision must match the Firmware revision in the device. To check it in PDM, go to SITRANS LUT400 HART > Identification > Device.

matching Firmware and EDD revisions

Installing a new version of SIMATIC PDM requires the most recent Service Pack (SP) and the most recent hot fix (HF).

To install a new EDD


Go to www.siemens.com/sitransLUT400 > Support > Software Downloads to download the most up-to-date EDD from the product page of our website. Save the files to your computer and extract the zipped file to an easily accessed location. Launch SIMATIC PDM Manage Device Catalog, browse to and select the folder which contains the unzipped EDD file. Page EN-32 SITRANS LUT400 (HART) 7ML19985XU81

Configuring a new device


1) 2) Check that you have the most recent EDD, and if necessary update it (seeTo install a new EDD, above). Configure the SITRANS LUT400 using the Quick Start Wizard for your application.

English

Quick Start Wizards via SIMATIC PDM


Notes:
The Quick Start Wizard settings are inter-related and changes apply only after you click on Apply and Transfer at the end of the final step. Click on BACK to return and revise a setting or CANCEL to exit the Quick Start. Open the Device menu in SIMATIC PDM and select the Quick Start for your application (Level, Volume, Volume-Linearization, Flow).

Quick Starts

Operating via AMS Device Manager


AMS Device Manager is a software package designed to commission and maintain SITRANS LUT400 and other process devices. You can find more information at: http://www.emersonprocess.com/AMS/.

Electronic Device Description (EDD)


SITRANS LUT400 requires the EDD for AMS Device Manager version 10.5.

Configuring a new device via AMS Device Manager


1) You can locate the EDD in Device Catalog, under Sensors/Level/Echo/Siemens AG/ SITRANS LUT400. Check the product page of our website at www.siemens.com/sitransLUT400, under Support->Software Downloads, to make sure you have the latest version of the EDD for AMS Device Manager. Save the files to your computer, and extract the zipped file to an easily accessed location. Launch AMS Device Manager Add Device Type, browse to the unzipped EDD file and select it. (An Application Guide for setting up HART devices with AMS Device Manager can be downloaded from the product page of our website under Support.) In Device Connection View, right-click on the device icon and select Scan Device to upload parameters from the device. Double click the device icon to open the startup screen. The startup screen shows device identification details, and a navigation window on the left-hand side of the screen. Configure the SITRANS LUT400 using the appropriate Quick Start Wizard.

2) 3)

4) 5) 6)

7ML19985XU81

SITRANS LUT400 (HART)

Page EN-33

Operating via FC375/475 (HART)


The FC375/FC475 HART Communicator is a handheld communication device that is easy to use, and provides universal support for other HART devices, such as the SITRANS LUT400. The menu structure of the FC375/475 is very similar to that of AMS Device Manager.

English

Electronic Device Description (EDD)


SITRANS LUT400 requires the EDD for FC375/FC475.

Configuring a new device via FC375/FC475


1) Check the product page of our website at www.siemens.com/sitransLUT400, under Support>Software Downloads, to make sure you have the latest version of the EDD for FC375/ FC475. Save the files to your computer, and extract the zipped file to an easily accessed location. To install the EDD, run Emersons Field Communicator Easy Upgrade Utility. Configure the SITRANS LUT400 using the Quick Start Wizard for your application.

2) 3) 4)

Operating via FDT (Field Device Tool)


FDT is a standard used in several software packages designed to commission and maintain field devices such as SITRANS LUT400. Two commercially available FDTs are PACTware and Fieldcare. FDT is very similar to PDM. To configure a field device via FDT you need the DTM (Device Type Manager) for the device. Siemens instruments use SITRANS DTM and an instrument EDD written for SITRANS DTM. 1) 2) First install SITRANS DTM on your system. You can download it from: http://www.siemens.com/sitransdtm. Click on Support then go to Software downloads. Install the SITRANS LUT400 (HART) EDD for SITRANS DTM. It can be downloaded from the product page of our website: Go to www.siemens.com/sitransLUT400 under Support and click on Software Downloads.

Configuring a new device via FDT


An Application Guide can be downloaded from the product page of our website under Support.

Service and Maintenance


SITRANS LUT400 requires no maintenance or cleaning under normal operating conditions.

Device Repair and Excluded Liability


For detailed information, please see the inside back cover.

Page EN-34

SITRANS LUT400 (HART)

7ML19985XU81

Wiring setups for hazardous area installations


The following wiring options are available for hazardous area installations: Dust Ignition Proof wiring (International/Europe) Non-incendive wiring (Canada) In all cases, check the device label on your instrument, and confirm the approval rating.

English

1. Dust Ignition Proof wiring (International/Europe)


SITRANS LUT420 7MLxxxx-xxxxx-xxxx Serial No.: GYZ / B1034567 0518 KCC-REM-S49 Power Rating: 100 230V 15% SITRANSLUT400 50/60 Hz, 36 VA (10 W) Contact Rating: 1A/5A @ 250V , Non-Inductive Operating Temperature: 20C to 50C Enclosure: IP65 / TYPE 4X / NEMA 4X II 3D Ex tc IIIC T70C Dc Ex tc IIIC T70C Dc SIRA 12ATEX9143X IECEx SIR 12.0057X Warning: Possible static hazard. Do not rub or clean on site with a dry cloth. Warning: Do not open when energized or when an explosive dust atmosphere is present. Siemens Milltronics Process Instruments, Peterborough Assembled in Canada with domestic and imported parts SITRANS LUT430 7MLxxxx-xxxxx-xxxx Serial No.: GYZ / B1034567 0518 KCC-REM-S49 Power Rating: 100 230V 15% SITRANSLUT400 50/60 Hz, 36 VA (10 W) Contact Rating: 1A/5A @ 250V , Non-Inductive Operating Temperature: 20C to 50C Enclosure: IP65 / TYPE 4X / NEMA 4X II 3D Ex tc IIIC T70C Dc Ex tc IIIC T70C Dc SIRA 12ATEX9143X IECEx SIR 12.0057X Warning: Possible static hazard. Do not rub or clean on site with a dry cloth. Warning: Do not open when energized or when an explosive dust atmosphere is present. Siemens Milltronics Process Instruments, Peterborough Assembled in Canada with domestic and imported parts SITRANS LUT440 7MLxxxx-xxxxx-xxxx Serial No.: GYZ / B1034567 0518 KCC-REM-S49 Power Rating: 100 230V 15% SITRANSLUT400 50/60 Hz, 36 VA (10 W) Contact Rating: 1A/5A @ 250V , Non-Inductive Operating Temperature: 20C to 50C Enclosure: IP65 / TYPE 4X / NEMA 4X II 3D Ex tc IIIC T70C Dc Ex tc IIIC T70C Dc SIRA 12ATEX9143X IECEx SIR 12.0057X Warning: Possible static hazard. Do not rub or clean on site with a dry cloth. Warning: Do not open when energized or when an explosive dust atmosphere is present. Siemens Milltronics Process Instruments, Peterborough Assembled in Canada with domestic and imported parts

SITRANS LUT420 7MLxxxx-xxxxx-xxxx Serial No.: GYZ / B1034567 0518 KCC-REM-S49 Power Rating: 10 32V , 10W SITRANSLUT400 Contact Rating: 1A/5A @ 250V , Non-Inductive Operating Temperature: 20C to 50C Enclosure: IP65 / TYPE 4X / NEMA 4X

SITRANS LUT430 7MLxxxx-xxxxx-xxxx Serial No.: GYZ / B1034567 0518 KCC-REM-S49 Power Rating: 10 32V , 10W SITRANSLUT400 Contact Rating: 1A/5A @ 250V , Non-Inductive Operating Temperature: 20C to 50C Enclosure: IP65 / TYPE 4X / NEMA 4X

SITRANS LUT440 7MLxxxx-xxxxx-xxxx Serial No.: GYZ / B1034567 0518 KCC-REM-S49 Power Rating: 10 32V , 10W SITRANSLUT400 Contact Rating: 1A/5A @ 250V , Non-Inductive Operating Temperature: 20C to 50C Enclosure: IP65 / TYPE 4X / NEMA 4X

II 3D Ex tc IIIC T70C Dc Ex tc IIIC T70C Dc SIRA 12ATEX9143X IECEx SIR 12.0057X Warning: Possible static hazard. Do not rub or clean on site with a dry cloth. Warning: Do not open when energized or when an explosive dust atmosphere is present.

II 3D Ex tc IIIC T70C Dc Ex tc IIIC T70C Dc SIRA 12ATEX9143X IECEx SIR 12.0057X Warning: Possible static hazard. Do not rub or clean on site with a dry cloth. Warning: Do not open when energized or when an explosive dust atmosphere is present.

II 3D Ex tc IIIC T70C Dc Ex tc IIIC T70C Dc SIRA 12ATEX9143X IECEx SIR 12.0057X Warning: Possible static hazard. Do not rub or clean on site with a dry cloth. Warning: Do not open when energized or when an explosive dust atmosphere is present.

Siemens Milltronics Process Instruments, Peterborough Assembled in Canada with domestic and imported parts

Siemens Milltronics Process Instruments, Peterborough Assembled in Canada with domestic and imported parts

Siemens Milltronics Process Instruments, Peterborough Assembled in Canada with domestic and imported parts

The ATEX certificate listed on the device label can be downloaded from the product page of our website at www.siemens.com/sitransLUT400. The IECEx certificate listed on the device label can be viewed on the IECEx website. Go to: http://iecex.iec.ch and click on Ex Equipment Certificates of Conformity then enter the certificate number IECEx SIR 12.0057X. For wiring requirements follow local regulations. See also the ATEX certificate listed above.

7ML19985XU81

SITRANS LUT400 (HART)

Page EN-35

2. Non-incendive wiring (Canada)


SITRANS LUT420 7MLxxxx-xxxxx-xxxx Serial No.: GYZ / B1034567 Power Rating: 100 230V 15% 50/60 Hz, 36 VA (10 W) Contact Rating: 1A/5A @ 250V , Non-Inductive Operating Temperature: 20C to 50C Enclosure: IP65 / TYPE 4X / NEMA 4X SITRANS LUT430 7MLxxxx-xxxxx-xxxx Serial No.: GYZ / B1034567 Power Rating: 100 230V 15% 50/60 Hz, 36 VA (10 W) Contact Rating: 1A/5A @ 250V , Non-Inductive Operating Temperature: 20C to 50C Enclosure: IP65 / TYPE 4X / NEMA 4X SITRANS LUT440 7MLxxxx-xxxxx-xxxx Serial No.: GYZ / B1034567 Power Rating: 100 230V 15% 50/60 Hz, 36 VA (10 W) Contact Rating: 1A/5A @ 250V , Non-Inductive Operating Temperature: 20C to 50C Enclosure: IP65 / TYPE 4X / NEMA 4X

English

159134

Class I, Div.2, Gr. A, B, C & D T3C Class II, Div. 2, Gr. F & G Class III Per Drawing A5E03936871

159134

Class I, Div.2, Gr. A, B, C & D T3C Class II, Div. 2, Gr. F & G Class III Per Drawing A5E03936871

159134

Class I, Div.2, Gr. A, B, C & D T3C Class II, Div. 2, Gr. F & G Class III Per Drawing A5E03936871

Siemens Milltronics Process Instruments, Peterborough Assembled in Canada with domestic and imported parts

Siemens Milltronics Process Instruments, Peterborough Assembled in Canada with domestic and imported parts

Siemens Milltronics Process Instruments, Peterborough Assembled in Canada with domestic and imported parts

SITRANS LUT420 7MLxxxx-xxxxx-xxxx Serial No.: GYZ / B1034567 Power Rating: 10 32V , 10W Contact Rating: 1A/5A @ 250V , Non-Inductive Operating Temperature: 20C to 50C Enclosure: IP65 / TYPE 4X / NEMA 4X

SITRANS LUT430 7MLxxxx-xxxxx-xxxx Serial No.: GYZ / B1034567 Power Rating: 10 32V , 10W Contact Rating: 1A/5A @ 250V , Non-Inductive Operating Temperature: 20C to 50C Enclosure: IP65 / TYPE 4X / NEMA 4X

SITRANS LUT440 7MLxxxx-xxxxx-xxxx Serial No.: GYZ / B1034567 Power Rating: 10 32V , 10W Contact Rating: 1A/5A @ 250V , Non-Inductive Operating Temperature: 20C to 50C Enclosure: IP65 / TYPE 4X / NEMA 4X

159134

Class I, Div.2, Gr. A, B, C & D T3C Class II, Div. 2, Gr. F & G Class III Per DWG. A5E03936871

159134

Class I, Div.2, Gr. A, B, C & D T3C Class II, Div. 2, Gr. F & G Class III Per DWG. A5E03936871

159134

Class I, Div.2, Gr. A, B, C & D T3C Class II, Div. 2, Gr. F & G Class III Per DWG. A5E03936871

Siemens Milltronics Process Instruments, Peterborough Assembled in Canada with domestic and imported parts

Siemens Milltronics Process Instruments, Peterborough Assembled in Canada with domestic and imported parts

Siemens Milltronics Process Instruments, Peterborough Assembled in Canada with domestic and imported parts

CSA Class I, Div 2 connection drawing number A5E03936871 can be downloaded from the product page of our website at www.siemens.com/sitransLUT400.

Instructions specific to hazardous area installations (Reference European ATEX Directive 94/9/EC, Annex II, 1/0/6)
The following instructions apply to equipment covered by certificate number SIRA 12ATEX9143X: 1) 2) 3) 4) For use and assembly, refer to the main instructions. The equipment is certified for use as Category 3D equipment. The equipment may be used in hazardous zone 22. This equipment has a maximum surface temperature of T70 C (in a +50 C ambient). Refer to the applicable Code Of Practice for selection of this equipment with respect to specific dust ignition temperatures. 5) The equipment is certified for use in an ambient temperature range of -20 to +50 C. 6) Any cable or conduit entries must meet the requirements of European Directive 94/9/EC for Group II, Category 3D and maintain the overall IP rating of the enclosure. 7) Cable should be selected in accordance with the applicable Code of Practice and such that its insulation can withstand the maximum surface temperature of the enclosure (T70 C). 8) The equipment has not been assessed as a safety related device (as referred to by Directive 94/9/EC Annex II, clause 1.5). 9) Installation and inspection of this equipment shall be carried out by suitably trained personnel in accordance with the applicable code of practice (EN 60079-14 and EN 6007917 in Europe). 10) Repair of this equipment shall be carried out by suitably trained personnel in accordance with the applicable code of practice (e.g. EN 60079-19 within Europe).

Page EN-36

SITRANS LUT400 (HART)

7ML19985XU81

11) Components to be incorporated into or used as replacements in the equipment shall be fitted by suitably trained personnel in accordance with the manufacturer's documentation. 12) Manual override can be accomplished by using the disconnect switch provided in the building installation.

English

SPECIAL CONDITIONS FOR SAFE USE


The 'X' suffix to the certificate number relates to the following special condition(s) for safe use: Any cable or conduit entries shall meet the requirements of the European Directive 94/9/EC for Group II, Category 3D and maintain the overall IP rating of the enclosure. Cable shall be selected in accordance with the applicable Code of Practice and its insulation shall be capable of withstanding the maximum surface temperature of the enclosure (T70 C). The enclosure is non-conducting and, under certain extreme conditions, may generate an ignition-capable level of electrostatic charges. The user shall ensure that the equipment is not installed in a location where it may be subjected to external conditions (such as high-pressure steam) which might cause a build-up of electrostatic charges on non-conducting surfaces. Additionally, cleaning of the equipment should be done only with a damp cloth. The equipment shall either be installed in an area where there is a low risk of mechanical damage or, if this is not possible, it shall be protected by an additional guard that provides it with mechanical protection that is adequate for the application. Any LUT400 series controller (LUT420, LUT430, LUT440) used with remote panel mounted lid shall be installed in a suitably rated enclosure meeting the requirements of the European Directive 94/9/EC for Group II, Category 3D and rated IP65 or better.

7ML19985XU81

SITRANS LUT400 (HART)

Page EN-37

Notes
English
Page EN-38

SITRANS LUT400 (HART)

7ML19985XU81

SITRANS LUT400 (HART) Kvikstart manual


Denne manual opridser de vsentligste karakteristika og funktioner af SITRANS LUT400 (HART 1)). Vi anbefaler Dem kraftigt at anskaffe den detaljerede version af denne manual, s De kan anvende apparatet fuldt ud. Den komplette SITRANS LUT400 (HART) betjeningsvejledning samt manualen Kommunikation for SITRANS LUT400 (HART), kan downloades fra produktsiden p vort website: www.siemens.com/sitransLUT400. Den trykte manual kan fs hos Deres lokale Siemens reprsentant. Sprgsml vedrrende indholdet af denne manual kan rettes til: Siemens AG Siemens Milltronics Process Instruments 1954 Technology Drive, P .O. Box 4225 Peterborough, Ontario, Canada, K9J 7B1 E-mail: techpubs.smpi@siemens.com

Dansk

Copyright Siemens AG 2012. Alle rettigheder forbeholdes


Vi opfordrer brugerne til at anskaffe de autoriserede, indbundne manualer eller at lse de elektroniske versioner, der er udarbejdet og skrevet af Siemens Milltronics Process Instruments. Siemens Milltronics Process Instruments ptager sig intet ansvar for indholdet af delvise eller fuldstndige gengivelser af indbundne eller elektroniske versioner.

Ansvarsfragelse
Sknt vi har kontrolleret, at indholdet af denne manual stemmer overens med de beskrevne instrumenter, kan der stadig forekomme variationer. Vi kan derfor ikke garantere en fuldstndig overensstemmelse. Indholdet af denne manual revideres jvnligt, og eventuelle rettelser inkluderes i de efterflgende udgaver. Vi modtager gerne forslag til forbedringer. Retten til ndringer af de tekniske data forbeholdes.

MILLTRONICS er et registreret varemrke, der tilhrer Siemens Milltronics Process Instruments.

Teknisk Support
Support er tilgngelig 24 timer i dgnet. Adresse, telefon- og faxnummer p Siemens Automations lokale kontor kan findes p: www.siemens.com/automation/partner: Klik p fanebladet Contact (Kontakt) for at finde den relevante produktgruppe (for eksempel: Automation Technology (Automatiseringsteknik)). Klik igen p produktgruppens dropdown-boks for at g til produktfamilien. For niveaumling: +Automation Technology (Automatiseringsteknik) > +Sensor Systems (Flersystemer) >+Process Instrumentation (Procesinstrumentering) > +Level Measurement (Niveaumling) > +Continuous (Kontinuert). Vlg Ultrasonic (Ultralyd). For flowmling: +Automation Technology > +Sensor Systems >+Process Instrumentation > +Flow Measurement (Flowmling). Vlg Ultrasonic. Vlg land og derefter by/egn. Vlg Technical Support (Teknisk support) under Service. For on-line teknisk support, g til: www.siemens.com/automation/support-request Indtast apparatets navn (SITRANS LUT400) eller ordrenummeret, klik p Search (Sg) og vlg den relevante produkttype. Klik p Next (Nste). Indtast et ngleord, der beskriver problemet. Gennemlb derefter den relevante dokumentation eller klik p Next for at e-maile en beskrivelse af problemet til personalet i Siemens Tekniske Support. Siemens IA/DT Teknisk Support Center: telefonnr. +49 (0)911 895 7222
1)

HARTer et registreret varemrke, der tilhrer HART Communication Foundation.

7ML19985XU81

SITRANS LUT400 (HART)

Side DA-1

Sikkerhedsvejledning
De anfrte advarsler skal overholdes for at sikre egen og andres sikkerhed samt for at beskytte produktet og det tilknyttede udstyr. Disse advarsler ledsages af en tydeliggrelse af graden af forsigtighed, der br overholdes. ADVARSEL: vedrrer et advarselssymbol p produktet og betyder, at en manglende overholdelse af de ndvendige forholdsregler kan fre til dd, alvorlig personskade og/eller omfattende materielle skader. ADVARSEL: anvendes, nr der ikke er noget tilsvarende advarselssymbol p produktet, og betyder, at en manglende overholdelse af de ndvendige forholdsregler kan fre til dd, alvorlig personskade og/eller omfattende materielle skader. Bemrk: betyder vigtige oplysninger om produktet eller denne del af betjeningsvejledningen.

Dansk

FCC-overensstemmelse
Kun for installationer i USA: Regler fra Federal Communications Commission (FCC)
ADVARSEL: ndringer eller modifikationer, der ikke er udtrykkeligt godkendt af Siemens Milltronics, kan ophve brugerens ret til at benytte udstyret. Bemrk: Dette udstyr er blevet testet og fundet at overholde grnserne for en klasse A digital anordning i henhold til Afsnit 15 i FCC-reglerne. Disse grnser er beregnet til at yde en rimelig beskyttelse mod skadelige interferenser, nr anordningen anvendes i kommercielle omgivelser. Dette udstyr frembringer, bruger og kan udstrle radiofrekvent energi, og kan, sfremt det ikke installeres og bruges i overensstemmelse med instruktionsbogen, forrsage interferens, der kan virke forstyrrende p radiokommunikation. Brugen af dette udstyr i et beboelsesomrde vil sandsynligvis forrsage skadelig interferens med radiokommunikationer, som brugeren i givet fald vil vre ndsaget til at udbedre for egen regning.

Side DA-2

SITRANS LUT400 (HART)

7ML19985XU81

SITRANS LUT400
ADVARSEL: SITRANS LUT400 br kun anvendes som beskrevet i denne manual, da den beskyttelse, udstyret yder, ellers kan forringes. Bemrk: Dette produkt er beregnet til at anvendes i industriomrder. Brugen at dette udstyr i et beboelsesomrde kan forrsage interferens med diverse frekvensbaserede kommunikationsmidler. Kontrolenhederne i Siemens SITRANS LUT400-serie er kompakte, enkeltpunkt, langtrkkende ultralydskontrolenheder til kontinuert niveaumling p vsker, opslmninger og faststoffer samt til njagtig overvgning af strmninger i ben kanal (OCM). Denne serie er kompatibel med det komplette udvalg af Siemens EchoMax transducere og tillader et driftsomrde fra 0,3 til 60 meter (afhngigt af transduceren). SITRANS LUT400 er forbundet med et Lokalt Brugerinterface (LBI) med baggrundsbelyst display, der giver adgang til menubaseret programmering. Instrumentet har et intuitivt, webbaseret interface samt en rkke plug-and-play-guider. SITRANS LUT400 understtter HART kommunikationsprotokol samt adskillige programpakker for fjernkommunikation [SIMATIC PDM, AMS Device Manager, Field Communicator 375/475 og Field Device Tools (FDTs)]. Det omfatter ogs nste generation af Sonic Intelligencefor endnu mere hjtydende mlinger og strre brugervenlighed.

Dansk

Tekniske data
Strmforsyning
AC-version

For en fuldstndig oversigt, se SITRANS LUT400 (HART) Betjeningsvejledning.

100-230 Vac Sikring: 5 x 20 mm, Slow Blow, 0,25A, 250V


Strmforbruget er angivet ved maksimum.

DC-version
10-32 Vdc, 10W 1 Sikring: 5 x 20 mm, Slow Blow, 1,6A, 125V

15%, 50 / 60 Hz, 36 VA (10W)1)

1)

Omgivelsesbetingelser
Hjde
max. 2.000 m

Omgivende temperatur
-20 til +50C (-4 til +122F)

Installationskategori
II

Forureningsgrad
4

Relativ luftfugtighed
Egnet til udendrsbrug (kun med IP65 / Type 4X / NEMA 4X indkapsling)

Godkendelser
Bemrk: Apparatets skilt angiver de godkendelser, der glder for netop Deres apparat. Brug passende kabelrrsttninger for at bevare IP- eller NEMA-klassificeringen. Almen brug Risikoomrder Ikke-antndingsfarlig (Canada) Stvantndelsesbeskyttet (Internationalt/Europa) CSAUSA/C, CE, FM, UL-certificeret, C-TICK CSA Klasse I, Div. 2, Gruppe A, B, C, D; Klasse II, Div 2, Gruppe F, G; Klasse III ATEX II 3D Ex tc IIIC T70 C IP65 IECEx SIR 12.0057X SIRA 12ATEX9143X Side DA-3

7ML19985XU81

SITRANS LUT400 (HART)

Installation
Bemrk: Installationen m kun foretages af kvalificeret personale og under overholdelse af de lokalt gldende bestemmelser. Dette produkt er flsomt over for elektrostatisk std. Flg derfor korrekte jordingsprocedurer. Alle feltledninger skal vre forsynet med tilstrkkelig isolering til mindst 250V. Der er farlige spndinger til stede p transducerterminalerne under drift. DC-indgangsklemmerne skal forsynes fra en kilde, der yder elektrisk isolation mellem ind- og udgang, for at opfylde de gldende sikkerhedskrav i henhold til IEC 61010-1. Relkontaktklemmerne er beregnet til brug med udstyr uden tilgngelige strmfrende dele og elinstallationer med tilstrkkelig isolation til mindst 250 V. Den maksimale tilladelige driftsspnding mellem nabo-relkontakter er p 250 V. Det ikke-metalliske hus giver ikke jording mellem rrforbindelserne. Brug jordende gennemfringer og bjler.

Dansk

Montageanvisning
Montageanvisningerne er forskellige for apparater, der monteres p vg, rr, DIN-skinne eller med fjerndisplay. Flg anvisningerne, der svarer til det aktuelle apparat.

Bemrk: Nr kabler fres gennem et kabelrr, skal Kabelfringsanvisningerne


p side 9 flges, inden SITRANS LUT400 monteres.

Vg- eller panelmontering


Alle konfigurationer af SITRANS LUT400 leveres med en monteringsbagplade. Modellen til panelmontering leveres med bde et LBI-display og et blankt dksel.
monteringsbagplade

hus med valgfrit displaydksel clips (2 steder) clips (2 steder)

Bemrk: Befstelseselementer til vgmontering medflger ikke.

Side DA-4

SITRANS LUT400 (HART)

7ML19985XU81

monteringsbagplade

hus med blankt dksel clips (2 steder) clips (2 steder)

Dansk

Bemrk: Befstelseselementer til vgmontering medflger ikke. husets bagside

elementer med slidser

Vgmontering af huset
1) 2) 3) Afmrk og bor fire huller i montageoverfladen til de fire skruer (medflger ikke). Skru dem fast med en skruetrkker. Hold apparatet, s slidserne p bagsiden str ud for clipsene p monteringsbagpladen. Tryk LUT400 mod bagpladen og skub nedad, s den sidder fast.

huller til monteringsskruer p bagpladen

Bemrk venligst:
Anbefalet strrelse af befstelseselementer: M8 eller 5/16" skrue med spndeskive med en maksimal udvendig diameter p 17 mm eller 5/8". Anbefalet montering: direkte montering p vggen. Hvis der anvendes en anden montageoverflade, SKAL denne vre i stand til at modst fire gange apparatets vgt.

7ML19985XU81

SITRANS LUT400 (HART)

Side DA-5

Fjernmonteret dksel

fjerndisplaydksel

pakning

Dansk

4 befstelseselementer til panelmontering medflger forlngerledning SITRANS LUT400

Montage af fjerndkslet
Bemrk: Det fjernmonterede dksel kan monteres op til 5 m fra apparatet ved hjlp af to forlngerledninger, som hver er 2,5 m lange (ekstraudstyr). Se den komplette Betjeningsvejledning. 1) Udskr det ndvendige hul til LBI-fjerndisplaydkslet ved hjlp af den medflgende skabelon. Anbring pakningen inden i dkslet svarende til monteringshullerne. Anbring bagsiden af fjerndisplaydkslet svarende til det udskrne hul i panelet. Afmrk og bor fire huller i montageoverfladen til de fire skruer (medflger). Skru dem fast med en skruetrkker og skruengle. Bemrk: Anbefalet moment p befstelsesskruerne for god tthed: 1,1 N m 10 in-lbs

2)

Bemrk venligst:
Anbefalet montering: montering p panel, op til 5 m fra apparatet. Hvis der anvendes en anden montageoverflade, SKAL denne vre i stand til at modst fire gange apparatets vgt. Bemrk: Medflgende befstelseselementer: M5 skrue, ttningsskive, M5 flad spndeskive og mtrik. Disse befstelseselementer er ndvendige for at bevare IP65-niveauet for det fjernmonterede dksel. Side DA-6 SITRANS LUT400 (HART) 7ML19985XU81

Rrmontering

monteringsbagplade rrbjle

bjleskrue

hus med valgfrit displaydksel

Dansk

rr Bemrk: Befstelseselementer til rrmontering medflger ikke.

Montering af huset
1) 2) 3) Fastgr monteringsbagpladen p rret ved hjlp af bjleskruer og rrbjler, der passer til rrets diameter (medflger ikke). Skru bjleskruerne fast med en skruengle. Pas p ikke at stramme s meget, at pladen bjes eller vrides. Det kan gre det umuligt at clipse LUT400 fast p bagpladen. Fastgr apparatet p monteringsbagpladen (som beskrevet i trin 3 af Vgmontering af huset p side 5).

Bemrk venligst:
Anbefalet montering: direkte p vandret eller lodret rr. Hvis der anvendes en anden montageoverflade, SKAL denne vre i stand til at modst fire gange apparatets vgt. Anbefalede rrdimensioner: maksimum: 3" rr, minimum: 3/4" rr Anbefalede strrelser af befstelseselementer: Bjleskrue maksimum: 3" rr med M8 eller 3/8" gevind minimum: 3/4" rr med M6 eller 1/4" gevind Sekskantmtrikker: M6 eller 1/4" til M8 eller 3/8" Spndeskive: maksimum: udvendig diameter 16 mm eller 13/16".

7ML19985XU81

SITRANS LUT400 (HART)

Side DA-7

Montering p DIN-skinne
husets bagside

DIN-skinne

clips (2)

bning

DIN-skinneglider

Dansk Montering af huset


1) 2) 3) Hold toppen af huset skrt i retning af DIN-skinnen og en smule over skinnens vre kant. Tryk huset nedad mod DIN-skinnen, s clipsene p husets bagside gr i hak med den vre kant af DIN-skinnen. Tryk apparatet mod DIN-skinnen, s DIN-skinneglideren gr i indgreb, hvilket sikrer, at huset sidder godt fast p DIN-skinnen.

Bemrk venligst:
Anbefalet montering: direkte p vandret DIN-skinne. Pkrvede dimensioner af DIN-skinnen: TH 35-7,5 eller TH 35-15 iflge standarden IEC 60715. DIN-skinnen SKAL vre i stand til at modst fire gange vgten af SITRANS LUT400.

Aftagning af huset
1) Fra apparatets forside, anbring en skruetrkker i bningen nederst p DINskinneglideren og lirk den nedad. Dette vil lsne glideren fra den nedre kant af DINskinnen. Hold glideren nede, samtidig med at huset skubbes opad for at lsne clipsene fra den vre kant af DIN-skinnen.

2)

Side DA-8

SITRANS LUT400 (HART)

7ML19985XU81

Forberedelse for kabelindfring


Kablerne kan fres gennem kabelrr, eller de kan indfres i huset gennem kabelbsninger. Flg frst trin 1-5 herunder for at afdkke kabelindfringshullerne, og flg dernst de resterende trin for brug med kabelrr eller kabelbsninger. 1) 2) 3) 4) 5) Kontroller, at husets dksel er lukket, og at fastgrelsesskruerne er lst fast. Anbring spidsen af en skruetrkker i rillen p den ydre omkreds af af dkfligen (se nedenstende illustration). Sl p enden af skruetrkkeren med hndfladen for at blotlgge indfringshullet. Lsn skruerne og tag dkslet af huset. Fjern plaststykkerne, der dkkede indfringshullerne, fra huset. Pas p ikke at beskadige de elektroniske komponenter med statisk elektricitet eller med det vrktj, der bruges til at blotlgge indfringshullerne.
SITRANS LUT400

Bemrk:
Nr dkfligen over kabelindfringshullet er fjernet, er indfringshullet 21,4 mm til 21,6 mm i diameter. M20 kabelbsninger (20 mm i diameter) og 1/2" NPT kabelrr (21,3 mm i diameter) passer til indfringshullet. Pas p, nr der vlges ttningselementer til indfringshullerne. Flade pakninger anbefales (snarere end O-ringe). Hvis der anvendes andre kabelbsninger, er det kundens ansvar at bevare IP65klassificeringen for indfringshullerne.

Dansk

kabelindfringshuller

dkflig (3 steder)

rille til skruetrkker (3 steder)

Kabel frt gennem kabelrr


(fortsat fra trin 1-5 herover) 6) Efter forberedelse for kabelindfring i trin 1-5 herover fastgres kabelrret til huset: brug kun fittings af passende strrelse, som er godkendt til vandttte applikationer. (Strrelsen af kabelrr er 1/2" NPT). 7) St huset dksel tilbage p plads og stram skruerne.

Kabel frilagt og indfrt gennem kabelbsningerne


(fortsat fra trin 1-5 herover) 6) Efter forberedelse for kabelindfring i trin 1-5 herover skrues bsningerne af og fastgres sikkert p huset. 7) Fr kablerne gennem bsningerne. Srg for at holde strmforsyningskablet adskilt fra signalkablerne, og forbind dernst kablerne med klemmerkkerne. 8) Stram bsningerne for at opn en god ttning. 9) St husets dksel tilbage p plads og stram skruerne.

7ML19985XU81

SITRANS LUT400 (HART)

Side DA-9

Tilslutning
Sikkerhedsanvisninger for tilslutning
For fuldstndige anvisninger om elinstallationen henvises der til den komplette Betjeningsvejledning. Kontrollr, at alle systemkomponenter er installeret i henhold til instruktionerne. Forbind alle kabelafskrmningerne til LUT400s afskrmningsklemmer (mrket med symbolet p apparatet). For at undg differentielle jordpotentialer skal det kontrolleres, at kabelafskrmningerne er korrekt jordforbundet. Hold de ubeskyttede ledere p afskrmede kabler s korte som muligt for at begrnse stjen p linjen som flge af parasittransmissioner og stjoptagelse.

Tilslutning af SITRANS LUT400


Dansk
ADVARSLER:

Kontrollr godkendelsesniveauet p instrumentets skilt. Brug passende kabelrrsttninger for at bevare IP- eller NEMAklassificeringen.
Bemrk: Separate kabler og kabelrr kan vre ndvendige for at overholde standardmetoderne for elinstallation eller de elektriske normer. Adgang til eltilslutningsrummet: 1) 2) 3) Lsn lseskruerne en 1/4 omgang. Lft dkslet op og mod venstre p hngslerne. Dkslet kan enten st bent, forbundet via hngslerne, eller det kan clipses af hngslerne og sttes til side for at f adgang til tilslutningsrummet.

4) 5) 6)

Udfr alle tilslutningerne iflge nedenstende anvisninger. Nr elinstallationen er afsluttet, sttes dkslet tilbage p apparatet. Stram lseskruerne.

Side DA-10

SITRANS LUT400 (HART)

7ML19985XU81

Tilslutningsrummet
Klemtavlen p LUT400 gr det muligt at forbinde alle ind- og udgange samtidigt. Klemmerkkerne kan tages ud for at lette tilslutningen. PAS P: Srg for, at klemmerkkerne termineres p det korrekte sted, nr de sttes p plads igen. Hvis det ikke er tilfldet, kan det medfre skader p apparatet eller p eksternt udstyr, der er forbundet med det. Bemrk: Anbefalet tilspndingsmoment for klemskruerne: 0,56 - 0,79 N m 5 - 7 in-lbs Pas p ikke at overspnde skruerne. Strmforsyning
TS-3 TRANSDUCER DISCRETE INPUTS INPUT POWER

strmforsyning

Dansk

8 1

9 2

10

11 0.25 A, 250 V SLOW BLOW

12 L1

13 L2/N 100 - 230 V

14

DC SYNC

SHLD

WHT

BLK

BLK

WHT

klemmerkker
mA OUT/HART BATTERY BR2032

batteri

RELAYS

DISPLAY 15 16 RLY 1 17 18 19 20 21 22 23 24

RLY 2

RLY 3 ACT

PAS SERVICE

USBforbindelse

displaykabel

mA HARTforbindelse

Strmforsyning
ADVARSLER: DC-indgangsklemmerne skal forsynes fra en kilde, der yder elektrisk isolation mellem ind- og udgang, for at opfylde de gldende sikkerhedskrav i henhold til IEC 61010-1. Alle felt-elinstallationer skal vre forsynet med en passende isolering i forhold til de nominelle spndinger. Vigtigt! Srg for at ethvert tilknyttet alarm/styringsudstyr er deaktiveret, inden SITRANS LUT400 tilsluttes strmforsyningen for frste gang, og indtil det er kontrolleret, at systemet fungerer tilfredsstillende.

7ML19985XU81

SITRANS LUT400 (HART)

Side DA-11

SITRANS LUT400 fs i modeller til veksel- eller jvnstrmsforsyning (AC eller DC).

Vekselstrm (AC)
INPUT POWER

Jvnstrm (DC)
INPUT POWER

12

13

14

12 10 - 32 V

13

14

L1 L2/N 100 - 230 V

AC: 100-230 V AC 15%, 50/60 Hz, 36 VA (10W)

DC: 10-32 V DC, 10W

Dansk

Bemrk: Srg for, at apparatet er forbundet med en plidelig jord. 1) 2) 3) For at tilslutte strmforsyningen afisoleres kablet p ca. 70 mm (2,75") fra enden af kablet, og ledningerne fres gennem kabelbsningen 1). Forbind ledningerne med klemmerne som vist: polariteten er angivet under klemmerkken. Jordforbind apparatet i henhold til de lokale bestemmelser. Bemrkninger vedrrende vekselstrmstilslutninger: Udstyret skal vre beskyttet af en 15 A sikring eller en strmafbryder i bygningens elinstallation. En strmafbryder eller kontakt i bygningens elinstallation, der markeres som afbryder, skal befinde sig i umiddelbar nrhed af udstyret og inden for operatrens rkkevidde.

Batteriet
SITRANS LUT400 leveres med et batteri allerede installeret. (Se illustration p side 11 for batteriets placering). Batteriet (BR2032) har en forventet levetid p ti r, og det pvirkes af den omgivende temperatur. Hvis strmforsyningen til LUT400 falder ud, vil batteriet gre det muligt at holde apparatets realtidsur krende, indtil strmmen genoprettes. (For udskiftning af batteriet, se den komplette Betjeningsvejledning).

1)Hvis kablerne fres gennem kabelrr, br der kun bruges muffer af en passende strrelse, der er

godkendt til vandttte applikationer.

Side DA-12

SITRANS LUT400 (HART)

7ML19985XU81

Transducere
Advarsel: Der er farlig spnding til stede p transducerterminalerne under driften. Fr transducerkablet i et jordet kabelrr af metal, adskilt fra de vrige elledninger (undtagen eventuelle ledninger til TS-3 temperaturfler). Bemrk: Brug ikke koaksialt kabel sammen med SITRANS LUT400 p grund af den hje spnding, der overfres p afskrmningen af koaksialt kabel LUT400s afskrmningsledning og den hvide transducerledning m ikke forbindes med hinanden: tilslut dem til separate klemmer Se bort fra ldre transducermanualer, der anbefaler en sdan fremgangsmde
TS-3 TRANSDUCER

Dansk

SHLD

BLK

WHT

Reler
Relkontakterne er vist i deres strmlse position. Alle relerne kan konfigureres som positiv eller negativ logik (se parameter 2.8.11.Relay Logic i den komplette Betjeningsvejledning).
RELAYS

BLK

WHT

Strmafbrydelse
Relerne 2 og 3 er normalt bne. Rel 1 kan forbindes, s det er enten normalt bent eller normalt lukket. I tilflde af afbrydelse af strmforsyningen vil relerne vende tilbage til deres normale tilstand.

15

16 RLY 1

17

18

19

20

21

RLY 2

RLY 3

Reldata
et Form C (NO eller NC) rel (rel 1), 1A ved 250 Vac, induktionsfrit, 3A ved 30 Vdc to Form A (NO) reler (rel 2 og 3), 5A ved 250 Vac, induktionsfri, 3A ved 30 Vdc

7ML19985XU81

SITRANS LUT400 (HART)

Side DA-13

HART-tilslutning mA-udgang (HART)


mA OUT/HART

For AKTIV HARTtilslutning (ved hjlp af LUT400s indbyggede strmforsyning) forbindes klemmerne 22 og 23.

22

23

24

ACT

PAS

For PASSIV HARTtilslutning (ved hjlp af ekstern strmforsyning) forbindes klemmerne 23 og 24.

For yderligere informationer henvises til mA-udgangsparametrene (2.5.Current Output) i oversigtsafsnittet om Parametre i Betjeningsvejledningen.

Dansk

Tilslutninger ved installationer i risikoomrder


Se side 35.

Idriftsttelse af SITRANS LUT400


Vi anbefaler, at applikationen konfigureres i flgende rkkeflge: Frst udfres den Kvikstartguide, der passer til Deres applikation (Niveau, Volumen, Flow). Dernst indstilles pumperne ved hjlp af Pumpestyringsguiden (hvis det er relevant). Til sidst konfigureres alarmerne eller andre kontrolenheder, sammentllere og samplere under henvisning til de forskellige parametre (se den komplette Betjeningsvejledning). Kvikstartguiderne kan startes p to mder: lokalt (se Kvikstartguider via LBI p side 17) ved fjernbetjening (se Andre Kvikstartguider: p side 17)

Aktivering af SITRANS LUT400


Tnd for apparatet. SITRANS LUT400 starter automatisk i Mlingsmode. En overgangsskrm viser frst Siemens logo, derefter vises den aktuelle firmware-version af den lokale brugerinterface (LBI), mens den frste mling behandles. Frste gang apparatet konfigureres, vil De blive bedt om at vlge et sprog (engelsk, tysk, fransk, spansk eller kinesisk). Den korrekte dato og tid skal vre indstillet, inden apparatet konfigureres. (For at ndre dem indstilles parameteren 2.14.Date and Time. Se den komplette Betjeningsvejledning for yderligere detaljer).

Side DA-14

SITRANS LUT400 (HART)

7ML19985XU81

LCD-display Display i Mlingsmode: Normal drift 1) Niveau


1 10
SITRANS LUT400 LEVEL

2
1/2

4 5

18.91
1

6
2

5 SAVING PARAMETERS

Flow
1 10 11 12
SITRANS LUT400 FLOW HEAD
1 2 3

3
1/2
L/S M

4 5 6

500313..
60.000

DAILY TOTAL,...

999999.99

5 SAVING PARAMETERS

1 navn 2 mlt vrdi (level (niveau), space (frirum), distance (afstand), volume (volumen), flow eller head (vskestand)) 3 vrdi, der vises [Primr Variabel (PV)=1 af 2, Sekundr Variabel (SV)=2 af 2] 4 indikator for skift1) mellem PV og SV 5 enheder 6 sjlediagram, der angiver niveauet 7 sekundrt omrde, der viser de konfigurerede reler (til venstre) og diskrete inputs (til hjre) 8 tekstzone, der viser statusmeddelelser 9 apparatets statusindikator 10 den valgte (primre) flermode: level, space, distance, volume, head eller flow 11 sekundr flermode = head, nr primr flermode = flow 12 sammentllingsvrdier: displayet skifter mellem den daglige sammentlling og den lbende sammentlling

Dansk

Ved fejl
0 LOE

8 tekstzone, der viser en fejlkode og en fejlmeddelelse 9 ikon for "service pkrvet" vises

Display i Programmode Navigationsskrm


aktuel menu / En synlig menusjle er tegn p, at parameterparameterTRANSDUCER 2.1.6 nummer menulisten er for lang til, at alle navn NO TRANSDUCER XRS-5 punkterne kan vises. XPS-10 bnd Bredden af bndet p menusjlen XPS-15 aktuelt XCT-8 punkt angiver menulistens lngde: et XCT-12 XPS-30 bredere bnd er tegn p frre menusjle punkter. Bndets position angiver det aktuelle punkts omtrentlige position i listen. Et bnd halvvejs nede af menusjlen angiver, at det aktuelle punkt er halvvejs nede af listen. Et bredere bnd er tegn p frre punkter.

1)

Tryk p

eller

for at omskifte.

7ML19985XU81

SITRANS LUT400 (HART)

Side DA-15

Parameterskrm
menu hvor parameteren findes
SENSOR
UNITS SENSOR MODE SENSOR MODE SECONDARY

Redigeringsskrm
2.1.6

parameternummer parametervrdi/ valg

TRANSDUCER
NO TRANSDUCER XPS-10 XPS-15 XCT-8 XCT-12 XPS-30

2.1.6

XRS-5

TRANSDUCER

parameternavn

FREQUENCY

XRS-5

PROGRAM mode
Ved hjlp af apparatets lokale trykknapper, tryk p menuniveau 1. Gennemlb menuen ved hjlp af , for at sl over i Programmode og bne , , .

Valg af en valgmulighed i en liste Dansk


a) Naviger hen til den nskede parameter. b) Tryk p HJRE pil for at bne Redigeringsmode. Det aktuelle valg fremhves.

c) Gennemlb valgmulighederne hen til et nyt valg. d) Tryk p HJRE pil for at godkende valget. Displayet vender tilbage til parameterskrmen og viser det nye valg.

ndring af en numerisk vrdi


a) Naviger hen til den nskede parameter, for eksempel Lavt kalibreringspunkt (2.2.1.). b) Efter valg af parameteren vises den aktuelle vrdi. c) Tryk p HJRE pil for at bne Redigeringsmode. Cursorens position er fremhvet. d) Brug VENSTRE og HJRE pil til at flytte cursoren hen p det tal, der skal ndres. e) Efterhnden som hvert tal fremhves (vlges), bruges OP skrue henholdsvis op og ned for tallet. f) Nr decimalpunktummet er valgt, bruges OP decimalpunktummet. og NED pil og NED pil til at flytte til at
LOW CALIB. PT. 2.2.1

ESC

+ _ _ _ _ _ _ 2. 0 M

OK

g) For at forlade en menu uden at gemme ndringerne holdes VENSTRE pil nedtrykket, indtil ESC fremhves. Tryk en gang til p VENSTRE pil for at forlade menuen uden at gemme ndringerne. Hvis den nye parametervrdi derimod er korrekt, holdes HJRE pil nedtrykket, indtil OK fremhves.

h) Tryk p HJRE pil for at godkende den nye vrdi. Displayet vender tilbage til parameterskrmen og viser det nye valg. Kontroller, at alt er korrekt.

Side DA-16

SITRANS LUT400 (HART)

7ML19985XU81

Kvikstartguider
For at konfigurere SITRANS LUT400 til applikationer med mling af niveau, volumen (standard beholderfaconer) eller flow bruges de nedenfor beskrevne Kvikstartguider via LBI. En guide for applikationer, der anvender mere komplekse beholderfaconer, der tilgngelig via SIMATIC PDM. Se Kvikstartguider via SIMATIC PDM p side 33.

Andre Kvikstartguider:
Der findes ogs andre Kvikstartguider, som anvender forskellige softwarepakker 1): SIMATIC PDM (HART) AMS (HART) FC375/475 (HART) FDTs (HART) Parameterkonfigurationstabeller (til at registrere data forud for konfiguration af apparatet vha. Kvikstartguiden) kan fs p vort website. G til www.siemens.com/sitransLUT400 > Support > Application Guides (Brugsanvisninger).

Dansk

Kvikstartguider via LBI


1) 2) Tryk p for at bne Programmode. Vlg Wizards (Guider) (1.), Quick Start (Kvikstart) (1.1.) og derefter den passende kvikstart: QS Level (Niveau-kvikstart) (1.1.1.), QS Volume (Volumen-kvikstart) (1.1.2.) eller QS Flow (Flow-kvikstart) (1.1.3.). [Flow-kvikstartguiden vises kun p LBI for modeller med konfigurationerne LUT430 (Pumpe og Flow) og LUT440 (OCM)]. Flg trinnene, og vlg dernst Finish (Afslut) for at gemme ndringerne i Kvikstartparametrene og vende tilbage til Programmenuen, og tryk s tre gange p vende tilbage til Mlingsmode. for at

3)

Bemrk: Kvikstartguidens indstillinger er indbyrdes forbundet og ndres frst, nr der vlges Finish i det sidste trin. Kvikstartguiden br ikke bruges til at ndre individuelle parametre: se i stedet afsnittet Parametre i den komplette Betjeningsvejledning. (Udfr frst de srlige indstillinger for den aktuelle applikation efter gennemfrelse af Kvikstarten). For yderligere information om parametre i guiderne, se den komplette Betjeningsvejledning.

1. Wizards (Guider) 1.1. Quick Start (Kvikstart) 1.1.1. QS Level (Niveau-kvikstart)


Brug denne guide til at konfigurere enkle niveauapplikationer. a) Tryk p HJRE pil for at aktivere PROGRAM mode og bne menuniveau 1: MAIN MENU (HOVEDMENU). b) Tryk to gange p HJRE pil c) Tryk p HJRE pil
1)

for at g til menupunkt 1.1.1.

for at bne QS Level (Niveau-kvikstart).

Se Manualen Kommunikation for SITRANS LUT400 (HART)

7ML19985XU81

SITRANS LUT400 (HART)

Side DA-17

d) P hvert trin trykkes p NED pilen for at acceptere standardvrdierne og g direkte til nste punkt, eller HJRE pil for at bne Redigeringsmode: det aktuelle valg fremhves. e) Gennemlb punkterne hen til det nskede punkt og tryk p HJRE pil for at gemme for at ndringen og derefter p NED pil fortstte. f) De kan nr som helst trykke p OP pil for at annullere guiden. Start p Niveau-kvikstartguide

QUICK START LEVEL


TRANSDUCER XRS-5
BACK CANCEL NEXT EDIT

QUICK START LEVEL


TRANSDUCER

NO TRANSDUCER XPS-10 XPS-15 XCT-8 XCT-12

XRS-5

for at g tilbage eller p VENSTRE pil

Dansk

Viser den guidetype, der gennemfres.


Valgmuligheder CANCEL (Annuller), START

Transducer

Angiver den Siemens transducer, der er tilsluttet apparatet.


NO TRANSDUCER (INGEN TRANSDUCER), XRS-5, XPS-10, XPS-15, XCT-8, Valgmuligheder XCT-12, XPS-30, XPS-40, XLT-30, XLT-60, STH Standardvrdi: NO TRANSDUCER

Definerer den type mling, der krves for applikationen (og den tilsvarende mAudgang).
flerens referencepunkt Hjt kalibreringspunkt frirum

Drift

afstand

niveau Lavt kalibreringspunkt

Uden for mleomrdet

Type
LEVEL (NIVEAU) SPACE (FRIRUM) DISTANCE (AFSTAND)

Beskrivelse
* Materialehjde

Referencepunkt
Lavt kalibreringspunkt (proces tom-niveau). Hjt kalibreringspunkt (proces fuld-niveau) Flerens referencepunkt (fortsttes p nste side)

Afstand til materialets overflade

Side DA-18

SITRANS LUT400 (HART)

7ML19985XU81

(fortsat fra foregende side)

Type

Beskrivelse
Br IKKE vlges. Hvis driftstypen angives som OTHER (Andet), betyder det, at apparatet er konfigureret som en niveaumler, men at det tidligere er indstillet til en anden type end LEVEL (Niveau), SPACE (Frirum) eller DISTANCE (Afstand). Driftstypen skal indstilles til LEVEL, SPACE eller DISTANCE for at kunne fortstte Niveau-kvikstartguiden.

Referencepunkt

OTHER (ANDET)

Temperaturkilde

Oprindelse af den temperatur, der anvendes til at justere lydens hastighed.


TRANSDUCER, FIXED TEMPERATURE (Fast temperatur), EXTERNAL TS-3 Valgmuligheder (Ekstern TS-3), AVERAGE OF SENSORS (Gennemsnit af flere) Standardvrdi: TRANSDUCER

Dansk

Fast Temperature

Brug denne funktion, hvis der ikke anvendes nogen temperaturflerenhed.

Vrdi

Interval: fra -100,0 til +150,0 C Standardvrdi: +20,0 C

Denne parameter vises kun, hvis FIXED TEMPERATURE (FAST TEMPERATUR) er valgt som Temperaturkilde. Enheder

Flerens mleenheder.
Valgmuligheder M, CM, MM, FT, IN Standardvrdi: M

Bemrk: I dette eksempel formodes alle vrdier at vre i meter (m).


Hjt kalibreringspunkt

Afstand fra Flerens referencepunkt til Hjt kalibreringspunkt: normalt proces fuldniveau.

Vrdi

Interval: 0,000 til 60,000 Standardvrdi: 0,000

Lavt kalibreringspunkt

Afstand fra Flerens referencepunkt til Lavt kalibreringspunkt: normalt proces tomniveau.

Vrdi

Interval: 0,000 til 60,000 Standardvrdi: 60,000

7ML19985XU81

SITRANS LUT400 (HART)

Side DA-19

Svarhastighed

Indstiller apparatets reaktionshastighed over for ndringer i mlingerne inden for mlomrdet.

Bemrk:

Svarhastigheden kan kun indstilles via Kvikstartguiden, og ndringer, der indfres i parametrene Fill Rate per Minute (2.3.1.), Empty Rate per Minute (2.3.2.) eller Damping Filter (2.3.3.) efter guidens gennemfrelse, vil trde i stedet for indstillingen af svarhastigheden. Svarhastighed vises altid i m/minut.
SLOW (Langsom) (0,1 M/MIN) Valgmuligheder MEDIUM (1,0 M/MIN) FAST (Hurtig) (10 M/MIN) Standardvrdi: SLOW (0,1 M/MIN)

Dansk

Brug en indstilling, der lige netop er hurtigere end den maksimale fyldnings- eller tmningshastighed (den strste af disse). Langsommere indstillinger giver hjere prcision, hurtigere indstillinger tillader hurtigere fluktuationer af niveauet. Afslutning af Niveau-kvikstartguide

For en vellykket Kvikstart skal alle ndringerne udfres.


BACK (Tilbage), CANCEL (Annuller), FINISH (Afslut) (Displayet vender tilbage til 1.1 Kvikstartmenu, nr Kvikstart er vel Valgmuligheder gennemfrt eller annulleres. Hvis CANCEL vlges, skrives ingen af ndringerne til apparatet).

For at overfre Kvikstartvrdierne til apparatet og vende tilbage til Programmenuen, tryk p NED pil (Finish (Afslut)). Tryk derefter tre gange p VENSTRE pil at vende tilbage til Mlingsmode. for

1.1.2. QS Volume (Volumen-kvikstart)


Brug denne guide til at konfigurere volumenapplikationer, der benytter standard beholderfaconer. a) Tryk p HJRE pil MENU. c) Tryk p NED pil for at aktivere PROGRAM mode og bne menuniveau 1: MAIN for at g til menupunkt 1.1.1. for at bne QS Volume (Volumen-kvikstart).
QUICK START VOLUME
TRANSDUCER XRS-5
BACK CANCEL NEXT EDIT

b) Tryk to gange p HJRE pil

og HJRE pil

d) P hvert trin trykkes p NED pilen for at acceptere standardvrdierne og g direkte til nste punkt, eller HJRE pil for at bne Redigeringsmode: det aktuelle valg fremhves. e) Gennemlb punkterne hen til det nskede punkt og tryk p HJRE pil p NED pil f) for at gemme ndringen og derefter for at fortstte. for at g for at annullere guiden.

QUICK START VOLUME


TRANSDUCER

NO TRANSDUCER XPS-10 XPS-15 XCT-8 XCT-12

XRS-5

De kan nr som helst trykke p OP pil tilbage eller p VENSTRE pil

Side DA-20

SITRANS LUT400 (HART)

7ML19985XU81

Start p Volumen-kvikstartguide

Viser den guidetype, der gennemfres.


Valgmuligheder CANCEL (Annuller), START

Transducer

Angiver den Siemens transducer, der er tilsluttet apparatet.


NO TRANSDUCER (INGEN TRANSDUCER), XRS-5, XPS-10, XPS-15, XCT-8, Valgmuligheder XCT-12, XPS-30, XPS-40, XLT-30, XLT-60, STH Standardvrdi: NO TRANSDUCER

Temperaturkilde

Oprindelse af den temperatur, der anvendes til at justere lydens hastighed.


TRANSDUCER, FIXED TEMPERATURE (Fast temperatur), EXTERNAL TS-3 Valgmuligheder (Ekstern TS-3), AVERAGE OF SENSORS (Gennemsnit af flere) Standardvrdi: TRANSDUCER

Dansk

Fast Temperature

Brug denne funktion, hvis der ikke anvendes nogen temperaturflerenhed.

Vrdi

Interval: fra -100,0 til +150,0 C Standardvrdi: +20,0 C

Denne parameter vises kun, hvis FIXED TEMPERATURE er valgt som Temperaturkilde. Beholderfacon

Angiver beholderens facon for at tillade SITRANS LUT400 at beregne volumen i stedet for niveau. Hvis der vlges NONE (Ingen) udfres der ingen omregning til volumen. Vlg den beholderfacon, der svarer til den overvgede beholder eller tank.
NONE (Ingen), LINEAR (Liner), CYLINDER (Cylinder), PARABOLIC BOTTOM (Parabolisk bund), HALF SPHERE BOTTOM (Halvsfrisk bund), FLAT SLOPED BOTTOM (Flad, hldende bund), PARABOLIC ENDS Valgmuligheder (Paraboliske ender), SPHERE (Sfre), CONICAL BOTTOM (Konisk bund), CURVE TABLE (Krum tabel), LINEAR TABLE (Liner tabel) Standardvrdi: LINEAR

Hvis der vlges CURVE TABLE (Krum tabel) eller LINEAR TABLE (Liner tabel), skal der indtastes vrdier for niveau- og volumenknkpunkter efter gennemfrelse af guiden. (Se parameter 2.6.7.Table 1-8 i den komplette Betjeningsvejledning). Enheder

Flerens mleenheder.
Valgmuligheder M, CM, MM, FT, IN Standardvrdi: M

Bemrk: I dette eksempel formodes alle vrdier at vre i meter (m).

7ML19985XU81

SITRANS LUT400 (HART)

Side DA-21

Hjt kalibreringspunkt

Afstand fra Flerens referencepunkt til Hjt kalibreringspunkt: normalt proces fuldniveau.

Vrdi

Interval: 0,000 til 60,000 Standardvrdi: 0,000

Lavt kalibreringspunkt

Afstand fra Flerens referencepunkt til Lavt kalibreringspunkt: normalt proces tomniveau.

Vrdi
Svarhastighed

Interval: 0,000 til 60,000 Standardvrdi: 60,000

Dansk

Indstiller apparatets reaktionshastighed over for ndringer i mlingerne inden for mlomrdet.

Bemrk:

Svarhastigheden kan kun indstilles via Kvikstartguiden, og ndringer, der indfres i parametrene Fill Rate per Minute (2.3.1.) eller Empty Rate per Minute (2.3.2.) efter guidens gennemfrelse, vil trde i stedet for indstillingen af svarhastigheden. Svarhastighed vises altid i m/minut.
SLOW (0,1 M/MIN) Valgmuligheder MEDIUM (1,0 M/MIN) FAST (10 M/MIN) Standardvrdi: SLOW (0,1 M/MIN)

Brug en indstilling, der lige netop er hurtigere end den maksimale fyldnings- eller tmningshastighed (den strste af disse). Langsommere indstillinger giver hjere prcision, hurtigere indstillinger tillader strre fluktuationer af niveauet. Dimension A

Hjden af beholderens bund, nr bunden er konisk, pyramidalsk, parabolisk, sfrisk eller flad, hldende.

Vrdi
Dimension L

Interval: 0,000 til 99,999 Standardvrdi: 0,000

Lngde af det cylindriske tvrsnit af en vandret beholder med parabolisk ende.

Vrdi
Volumenenheder

Interval: 0,000 til 99,999 Standardvrdi: 0,000

Angiver enhederne for volumenmling.


Valgmuligheder L, USGAL, IMPGAL, CUM, USER DEFINED * (brugerdefineret) Standardvrdi: L

* Hvis der vlges USER DEFINED (brugerdefineret), skal vrdien indstilles efter gennemfrelse af guiden. Se parameter 2.6.6.User Defined Unit i den komplette Betjeningsvejledning. Side DA-22 SITRANS LUT400 (HART) 7ML19985XU81

Maksimalt volumen

Beholderens maksimale volumen. Angiv beholderens volumen svarende til Hjt kalibreringspunkt. Hvis for eksempel beholderens maksimale volumen er 8000 L, indtastes vrdien 8000.

Vrdi

Interval: 0,0 til 9999999 Standardvrdi: 100,0

Afslutning af Volumen-kvikstartguide

For en vellykket Kvikstart skal alle ndringerne udfres.


Valgmuligheder BACK (Tilbage), CANCEL (Annuller), FINISH (Afslut) (Skrmen vender tilbage til Kvikstartmenu 1.1, nr Kvikstarten er vel gennemfrt eller annulleres. Hvis CANCEL vlges, skrives ingen af ndringerne til apparatet).

Dansk

For at overfre Kvikstartvrdierne til apparatet og vende tilbage til Programmenuen, tryk p NED pil (Finish (Afslut)). Tryk derefter tre gange p VENSTRE pil at vende tilbage til Mlingsmode. for

1.1.3. QS Flow (Flow-kvikstart)


Brug denne guide til at konfigurere enkle flowapplikationer. [Vises kun p modeller med konfigurationerne LUT430 (Pumpe og Flow) og LUT440 (OCM)]. a) Tryk p HJRE pil MENU. for at aktivere PROGRAM mode og bne menuniveau 1: MAIN for at g til menupunkt 1.1.1. og HJRE pil for at bne QS Flow (Flow-kvikstart).
QUICK START FLOW
TRANSDUCER XRS-5
BACK CANCEL NEXT EDIT

b) Tryk to gange p HJRE pil c) Tryk to gange p NED pil

d) P hvert trin trykkes p NED pilen for at acceptere standardvrdierne og g direkte til nste punkt, eller HJRE pil for at bne Redigeringsmode: det aktuelle valg fremhves. e) Gennemlb punkterne hen til det nskede punkt og tryk p HJRE pil for at gemme for at ndringen og derefter p NED pil fortstte. f) De kan nr som helst trykke p OP pil for at annullere guiden. Start p Flow-kvikstartguide

QUICK START FLOW


TRANSDUCER

NO TRANSDUCER XPS-10 XPS-15 XCT-8 XCT-12

XRS-5

for at g tilbage eller p VENSTRE pil

Viser den guidetype, der gennemfres.


Valgmuligheder CANCEL (Annuller), START

7ML19985XU81

SITRANS LUT400 (HART)

Side DA-23

Transducer

Angiver den Siemens transducer, der er tilsluttet apparatet.


NO TRANSDUCER (INGEN TRANSDUCER), XRS-5, XPS-10, XPS-15, XCT-8, Valgmuligheder XCT-12, XPS-30, XPS-40, XLT-30, XLT-60, STH Standardvrdi: NO TRANSDUCER

Temperaturkilde

Oprindelse af den temperatur, der anvendes til at justere lydens hastighed.


TRANSDUCER, FIXED TEMPERATURE (Fast temperatur), EXTERNAL TS-3 Valgmuligheder (Ekstern TS-3), AVERAGE OF SENSORS (Gennemsnit af flere) Standardvrdi: TRANSDUCER

Fast Temperature

Oprindelse af den temperatur, der anvendes til at justere lydens hastighed.

Dansk

Vrdi

Interval: fra -100,0 til +150,0 C Standardvrdi: +20,0 C

Denne parameter vises kun, hvis FIXED TEMPERATURE er valgt som Temperaturkilde. Primr mleanordning

Angiver den primre mleanordning (PMD), der skal anvendes i applikationen.


EXPONENTIAL DEVICES (Eksponentielle anordninger), RECTANGULAR FLUME BS-3680 (Rektangulr mlekanal), ROUND NOSE HORIZONTAL CR. BS-3680 (Rundnset stemmevrk med vandret overfald), TRAPEZOIDAL FLUME BS-3680 (Trapezformet mlekanal), U-FLUME BS-3680 (U-formet mlekanal), FINITE CREST WEIR BS-3680 (Stemmevrk med begrnset overfald), THIN PLATE RECT. WEIR BS-3680 (Tyndpladet rektangulrt Valgmuligheder stemmevrk), THIN PLATE V-NOTCH WEIR BS-3680 (Tyndpladet stemmevrk med V-overfald), RECT. WEIR CONTRACTED (Rektangulrt stemmevrk med indsnvring), ROUND PIPE (Rundt rr), PALMER BOWLUS FLUME (Palmer Bowlus mlekanal), H-FLUME (H-mlekanal), OTHER (Andet)* Standardvrdi: EXPONENTIAL DEVICES

* Dette valg vil vre sat til OTHER (Andet), hvis guiden tidligere er udfrt via et HART software-vrktj (ssom SIMATIC PDM), og apparatet var sat til OFF eller UNIVERSAL HEAD VS. FLOW (Universel vandstand som funktion af flow). Hvis dette er udgangskonfigurationen, kan den primre mleanordning kun indstilles for Ingen beregning (OFF) eller for Linearisering (UNIVERSAL HEAD VS. FLOW) via HART software-vrktjer (SIMATIC PDM, AMS, FC375/475). Se den komplette Betjeningsvejledning. Flowberegningsmetode

Indstiller flowberegningsmetoden.
Valgmuligheder ABSOLUTE (Absolut), RATIOMETRIC (Ratiometrisk) Standardvrdi: ABSOLUTE

Side DA-24

SITRANS LUT400 (HART)

7ML19985XU81

Enheder

Flerens mleenheder.
Valgmuligheder M, CM, MM, FT, IN Standardvrdi: M

Bemrk: I dette eksempel formodes alle vrdier at vre i meter (m).


Hjt kalibreringspunkt

Afstand fra Flerens referencepunkt til Hjt kalibreringspunkt: normalt proces fuldniveau.

Vrdi

Interval: 0,000 til 60,000 Standardvrdi: 0,000

Lavt kalibreringspunkt

Afstand fra Flerens referencepunkt til Lavt kalibreringspunkt: normalt proces tomniveau.

Dansk

Vrdi
Svarhastighed

Interval: 0,000 til 60,000 Standardvrdi: 60,000

Indstiller apparatets reaktionshastighed over for ndringer i mlingerne inden for mlomrdet.

Bemrk:

Svarhastigheden kan kun indstilles via Kvikstartguiden, og ndringer, der indfres i parametrene Fill Rate per Minute (2.3.1.) eller Empty Rate per Minute (2.3.2.) efter guidens gennemfrelse, vil trde i stedet for indstillingen af svarhastigheden. Svarhastighed vises altid i m/minut.
SLOW (0,1 M/MIN) Valgmuligheder MEDIUM (1,0 M/MIN) FAST (10 M/MIN) Standardvrdi: SLOW (0,1 M/MIN)

Brug en indstilling, der lige netop er hurtigere end den maksimale fyldnings- eller tmningshastighed (den strste af disse). Langsommere indstillinger giver hjere prcision, hurtigere indstillinger tillader strre fluktuationer af niveauet. Beregningsfaktorer:

Bemrk:

Flgende fem parametre vises i guiden alt efter den ovenfor valgte primre mleanordning. Disse parametre anvendes i flowberegningsformlen (se den komplette Betjeningsvejledning for yderligere detaljer). Floweksponent (Primr mleanordning = EXPONENTIAL DEVICES (Eksponentielle anordninger))

Eksponenten til flowberegningsformlen.

Vrdi
7ML19985XU81

Interval: -999,000 til 9999,000 Standardvrdi: 1,550

SITRANS LUT400 (HART)

Side DA-25

K-faktor (Primr mleanordning = EXPONENTIAL DEVICES (Eksponentielle anordninger))

Konstant, der kun anvendes i flowberegningsformlen til absolut beregning af en eksponentiel anordning.

Vrdi
V-hakkets vinkel

Interval: -999,000 til 9999,000 Standardvrdi: 1,000

(Primr mleanordning = THIN PLATE V-NOTCH WEIR - Tyndpladet stemmevrk med V-overfald)

Vinkel af V-hakket, der anvendes i flowberegningsformlen.

Vrdi Dansk
Hldning

Interval: 25,000 til 95,000 Standardvrdi: 25,000

(Primr mleanordning = TRAPEZOIDAL FLUME eller ROUND PIPE - Trapezformet mlekanal eller Rundt rr)

Flowhldning, der anvendes i flowberegningsformlen.

Vrdi

Interval: -999,000 til 9999,000 Standardvrdi: 0,000

Ruhedskoefficient (Primr mleanordning = ROUND PIPE - Rundt rr)

Ruhedskoefficient, der anvendes i flowberegningsformlen.

Vrdi

Interval: -999,000 til 9999,000 Standardvrdi: 0,000

Side DA-26

SITRANS LUT400 (HART)

7ML19985XU81

Dimensioner af den primre mleanordning

Bemrk:

For hver primr mleanordning p nr Exponential Devices (Eksponentielle anordninger) og Other (Andet) skal der angives op til fire dimensioner. Guiden beder om hver dimension, der er ndvendig for den valgte primre mleanordning, og navnene p de forskellige dimensioner af den primre mleanordning vises. Primr mleanordning Dimensionens navn i guiden (reference i parametermenuen)
APPROACH WIDTH B (Tillbsbredde) (2.15.4.5. OCM Dimension 1) THROAT WIDTH B (Kanalbredde) (2.15.4.6. OCM Dimension 2)

Rectangular Flume BS-3680 (Rektangulr mlekanal)

Dansk

HUMP HEIGHT P (Pukkelhjde) (2.15.4.7. OCM Dimension 3) THROAT LENGTH L (Kanallngde) (2.15.4.8. OCM Dimension 4)

Round Nose Horizontal Crest Weir BS-3680 (Rundnset stemmevrk med vandret overfald)
CREST WIDTH B (Overfaldsbredde) (2.15.4.5. OCM Dimension 1) CREST HEIGHT P (Overfaldshjde) (2.15.4.6. OCM Dimension 2) CREST LENGTH L (Overfaldslngde) (2.15.4.7. OCM Dimension 3)

Trapezoidal Flume BS-3680


APPROACH WIDTH B (Tillbsbredde) (2.15.4.5. OCM Dimension 1) THROAT WIDTH B (Kanalbredde) (2.15.4.6. OCM Dimension 2) HUMP HEIGHT P (Pukkelhjde) (2.15.4.7. OCM Dimension 3) THROAT LENGTH L (Kanallngde) (2.15.4.8. OCM Dimension 4)

U-Flume BS-3680
APPROACH DIAMETER DA (Tillbsdiameter) (2.15.4.5. OCM Dimension 1) THROAT DIAMETER D (Kanaldiameter) (2.15.4.6. OCM Dimension 2) HUMP HEIGHT P (Pukkelhjde) (2.15.4.7. OCM Dimension 3) THROAT LENGTH L (Kanallngde) (2.15.4.8. OCM Dimension 4)

Finite Crest Weir BS-3680 (Stemmevrk med begrnset overfald)


CREST WIDTH B (Overfaldsbredde) (2.15.4.5. OCM Dimension 1) CREST HEIGHT P (Overfaldshjde) (2.15.4.6. OCM Dimension 2) CREST LENGTH L (Overfaldslngde) (2.15.4.7. OCM Dimension 3)

Thin Plate Rectangular Weir BS-3680 (Tyndpladet rektangulrt stemmevrk)


APPROACH WIDTH B (Tillbsbredde) (2.15.4.5. OCM Dimension 1) CREST WIDTH B (Overfaldsbredde) (2.15.4.6. OCM Dimension 2) CREST HEIGHT P (Overfaldshjde) (2.15.4.7. OCM Dimension 3)

Rectangular Weir Contracted (Rektangulrt stemmevrk med indsnvring)


CREST WIDTH B (Overfaldsbredde) (2.15.4.5. OCM Dimension 1) (fortsttes p nste side)

7ML19985XU81

SITRANS LUT400 (HART)

Side DA-27

Primr mleanordning (fortsat) Round Pipe (Rundt rr)

Dimensionens navn i guiden (reference i parametermenuen)


PIPE INSIDE DIAMETER D (Indvendig rrdiameter) (2.15.4.5. OCM Dimension 1)

Palmer Bowlus Flume (Palmer Bowlus mlekanal)


MAXIMUM FLUME WIDTH HMAX (Maksimal kanalbredde) (2.15.4.5. OCM Dimension 1)

H-Flume (H-mlekanal)
MAXIMUM LISTED HEAD HMAX (Maksimal registreret vskestand) (2.15.4.5. OCM Dimension 1)

Maksimal vskestand

Dansk

Det maksimale niveau knyttet til denne primre mleanordning.

Vrdi

Interval: 0,000 til 60,000 Standardvrdi: 60,000

Offset ved vskestand nul

Forskellen (positiv eller negativ) mellem Lavt kalibreringspunkt og vskestand nul (niveau ved nul-flow).

Vrdi

Interval: -60,000 til 60,000 Standardvrdi: 0,000

Enheder for strmningshastigheden

Volumenenhed, der bruges til visning af totalt flow.


L/S, L/MIN, CUFT/S, CUFT/D, GAL/MIN, GAL/D, IMPGAL/MIN, IMPGAL/D, Valgmuligheder CUM/H, CUM/D, USER DEFINED (Brugerdefineret)* Standardvrdi: L/S

* Hvis der vlges USER DEFINED (Brugerdefineret), skal vrdien indstilles efter gennemfrelse af guiden. Se parameter 2.6.6.User Defined Unit i den komplette Betjeningsvejledning. Maksimalt flow ved 20 mA

Den maksimale strmningshastighed.

Vrdi

Interval: -999 til 9999999 Standardvrdi: 100

Decimaler for strmningshastighed

Det maksimale antal decimaler, der skal vises.


Valgmuligheder NO DIGITS (Ingen tal), 1 DIGIT (1 tal), 2 DIGITS (2 tal), 3 DIGITS (3 tal) Standardvrdi: NO DIGITS

Afskring for lavt flow

Undlader sammentlling for niveauer af vskestand lig med eller under afskringsvrdien.

Vrdi

Interval: 0,000 til 60,000 Standardvrdi: 0,000

Side DA-28

SITRANS LUT400 (HART)

7ML19985XU81

Afslutning af Flow-kvikstartguide

For en vellykket Kvikstart skal alle ndringerne udfres.


Valgmuligheder BACK (Tilbage), CANCEL (Annuller), FINISH (Afslut) (Skrmen vender tilbage til Kvikstartmenu 1.1.1, nr Kvikstarten er vel gennemfrt eller annulleres. Hvis CANCEL vlges, skrives ingen af ndringerne til apparatet).

For at overfre Kvikstartvrdierne til apparatet og vende tilbage til Programmenuen, (Finish (Afslut)). Tryk derefter tre gange p VENSTRE pil tryk p NED pil at vende tilbage til Mlingsmode. for

Bemrk: Det anbefales kraftigt at udfre en 2.15.2.Auto Zero Head efter

gennemfrelse af guiden for at sikre den bedst mulige njagtighed. Se den komplette Betjeningsvejledning for yderligere detaljer.

Dansk

1.2. Pumpestyring
Brug denne guide til at konfigurere pumper, som skal anvendes i applikationen. Srg for frst at gennemfre den relevante Kvikstartguide. a) Tryk p HJRE pil MENU. b) Tryk p HJRE pil c) Tryk p NED pil for at aktivere PROGRAM mode og bne menuniveau 1: MAIN for at g til menupunkt 1.1. og HJRE pil for at bne Pump Control (Pumpestyring). for at acceptere standardvrdierne og g

d) P hvert trin trykkes p NED pilen direkte til nste punkt, eller HJRE pil

for at bne Redigeringsmode: det aktuelle valg fremhves. for at

e) Gennemlb punkterne hen til det nskede punkt og tryk p HJRE pil gemme ndringen og derefter p NED pil f) De kan nr som helst trykke p OP pil for at annullere guiden. Start af guide - Pumpestyring for at fortstte.

for at g tilbage eller p VENSTRE pil

Viser den guidetype, der gennemfres.


Valgmuligheder CANCEL (Annuller), START

Antal pumper

Vlg antallet af pumper, der skal anvendes med pumpestyring.


Valgmuligheder NONE (Ingen), 2 Standardvrdi: NONE

Hvis vrdien er sat til NONE, er pumpestyring slet fra.

7ML19985XU81

SITRANS LUT400 (HART)

Side DA-29

Rel for Pumpe 1

Vlger det rel, der svarer til Pumpe 1.


Valgmuligheder RELAY 2 (rel 2), RELAY 3 (rel 3) Standardvrdi: RELAY 2

Rel for Pumpe 2

Kan ikke redigeres. Angiver automatisk relet, der svarer til Pumpe 2, ud fra det rel, der er valgt for Pumpe 1 i det foregende trin.
Valgmuligheder Hvis Rel for Pumpe 1 = RELAY 2, s er Rel for Pumpe 2 = RELAY 3 (kun visning) Hvis Rel for Pumpe 1 = RELAY 3, s er Rel for Pumpe 2 = RELAY 2

Pumpestyringsmode

Indstiller styringsalgoritmen, der bruges til at sl relet til og fra.

Dansk

LUT420 Niveaustyring: ALTERNATE DUTY ASSIST, ALTERNATE DUTY BACKUP LUT430 Pumpe- og flowstyring: ALTERNATE DUTY ASSIST, ALTERNATE DUTY BACKUP , SERVICE RATIO , FIXED DUTY ASSIST, FIXED Valgmuligheder DUTY ASSIST, SERVICE RATIO DUTY BACKUP for forskellige DUTY BACKUP modeller LUT440 Hjprcisionsmling i ben kanal (OCM): ALTERNATE DUTY ASSIST, ALTERNATE DUTY BACKUP , SERVICE RATIO DUTY ASSIST, SERVICE RATIO DUTY BACKUP , FIXED DUTY ASSIST, FIXED DUTY BACKUP Standardvrdi (alle modeller): ALTERNATE DUTY ASSIST

Driftstidsforhold Pumpe 1

Vlger brug af pumpe baseret p driftstidsforholdet snarere end p den senest brugte.

Vrdi

Interval: 0 til 255 Standardvrdi: 1

Denne parameter vises kun, hvis en Service Ratio (Driftstidsforhold)-algoritme er valgt som Pumpestyringsmode. Driftstidsforhold Pumpe 2

Vlger brug af pumpe baseret p driftstidsforholdet snarere end p den senest brugte.

Vrdi

Interval: 0 til 255 Standardvrdi: 1

Denne parameter vises kun, hvis en Service Ratio (Driftstidsforhold)-algoritme er valgt som Pumpestyringsmode.

Side DA-30

SITRANS LUT400 (HART)

7ML19985XU81

Driftstid Rel 2

Indstiller hvor lang tid, i timer, pumperel 2 har krt.

Vrdi

Interval: 0 til 999999 Standardvrdi: 0

Brug standardvrdien for nye pumper eller indstil denne vrdi for eksisterende pumper med akkumuleret driftstid. (Denne parameter vises kun, hvis en Service Ratio (Driftstidsforhold)-algoritme er valgt som Pumpestyringsmode). Driftstid Rel 3

Indstiller hvor lang tid, i timer, pumperel 3 har krt.

Vrdi

Interval: 0 til 999999 Standardvrdi: 0

Brug standardvrdien for nye pumper eller indstil denne vrdi for eksisterende pumper med akkumuleret driftstid. (Denne parameter vises kun, hvis en Service Ratio (Driftstidsforhold)-algoritme er valgt som Pumpestyringsmode). ON-referencepunkt Pumpe 1

Dansk

Niveau, hvor Pumpe 1 tndes (ON), som defineret i 2.1.1.Units.

Vrdi

Interval: 0,000 til 99999,000 Standardvrdi: 0,000

ON-referencepunkt Pumpe 2

Niveau, hvor Pumpe 2 tndes (ON), som defineret i 2.1.1.Units.

Vrdi

Interval: 0,000 til 99999,000 Standardvrdi: 0,000

OFF referencepunkt Pumpe 1

Niveau, hvor Pumpe 1 slukkes (OFF), som defineret i 2.1.1.Units.

Vrdi

Interval: 0,000 til 99999,000 Standardvrdi: 0,000

OFF referencepunkt Pumpe 2

Niveau, hvor Pumpe 2 slukkes (OFF), som defineret i 2.1.1.Units.

Vrdi

Interval: 0,000 til 99999,000 Standardvrdi: 0,000

Afslutning af guide - Pumpestyring

For en vellykket guide skal alle ndringerne udfres.

Valgmulig BACK (Tilbage), CANCEL (Annuller), FINISH (Afslut) (Skrmen vender tilbage til Pumpestyringsmenuen, nr guiden er vel gennemfrt eller annulleres. Hvis heder CANCEL vlges, skrives ingen af ndringerne til apparatet).
For at overfre vrdierne til apparatet og vende tilbage til Programmenuen, tryk p NED pil (Finish (Afslut)). Tryk derefter to gange p VENSTRE pil tilbage til Mlingsmode. for at vende

7ML19985XU81

SITRANS LUT400 (HART)

Side DA-31

SITRANS LUT400 Kommunikation: HART


SITRANS LUT400 kan konfigureres via HART-netvrket ved hjlp af Emersons HART Communicator 375/475 (se den komplette Betjeningsvejledning for yderligere detaljer) eller ved hjlp af en softwarepakke. Den anbefalede softwarepakke er Siemens SIMATIC Process Device Manager (PDM). Detaljerne er beskrevet i brugsanvisningen Working with HART, som kan downloades fra produktsiden p vort website. G til: www.siemens.com/sitransLUT400 under Support , og klik p Application Guides (Brugsanvisninger).

Fjernbetjening
SITRANS LUT400 understtter flere softwarevrktjer til betjening via fjernkommunikation: PC, der krer SIMATIC PDM PC, der krer Emerson AMS Device Manager PC, der krer en webbrowser PC, der krer et Field Device Tool (FDT) Field Communicator 375/475 (FC375/FC475). Herunder er givet de vigtigste informationer for at kunne bruge disse vrktjer sammen med SITRANS LUT400. For hvert vrktj fs yderligere detaljer i manualen Kommunikation for SITRANS LUT400 (HART).

Dansk

SIMATIC PDM
SIMATIC PDM er en softwarepakke, der bruges til at idriftstte og vedligeholde SITRANS LUT400 og andre procesapparater. Flere oplysninger fs p www.siemens.com/simatic-pdm. Tjek supportsiden p vort website for at sikre Dem, at De har den nyeste version af SIMATIC PDM, den seneste Service Pack (SP) og den seneste hot fix (HF). G til: http://support.automation.siemens.com/WW/ llisapi.dll?func=cslib.csinfo&lang=en&objiD=10806857&subtype=133100

Opdatering af Electronic Device Description (EDD - Elektronisk apparatbeskrivelse)


EDD kan findes i Apparatkataloget (Device Catalog) under Sensors/Level/Echo/Siemens AG/ SITRANS LUT400 (Flere/Niveau/Ekko/Siemens AG/SITRANS LUT400). (Denne EDD er udarbejdet under hensyntagen til fremadkompatibilitet). En retningslinje til at vlge den korrekte EDD er, at hoved- og binumrene skal stemme overens for EDD-versionen og versionen af firmware i apparatet (eksempel p hoved- og binumre med fed skrift: 1.00.00-02).
EDD-versionen skal svare til versionen af firmware i apparatet. Dette tjekkes i PDM ved at g til SITRANS LUT400 HART > Identification (Identifikation) > Device (Apparat) .

matchende versioner af firmware og EDD

Installation af en ny version af SIMATIC PDM krver den nyeste Service Pack (SP) og den nyeste hot fix (HF).

For at installere en ny EDD


G til www.siemens.com/sitransLUT400 > Support > Software Downloads for at downloade den nyeste EDD fra produktsiden p vort website. Gem filerne p computeren og pak den zippede fil ud til en lettilgngelig placering. Start SIMATIC PDM Manage Device Catalog (Administrere Apparatkatalog), find den mappe, der indeholder den udpakkede EDD-fil, og vlg den. Side DA-32 SITRANS LUT400 (HART) 7ML19985XU81

Konfiguration af et nyt apparat


1) 2) Tjek, at De rder over den seneste EDD, og opdater den om ndvendigt (seFor at installere en ny EDD herover). Konfigurer SITRANS LUT400 ved hjlp af Kvikstartguiden, der passer til applikationen.

Kvikstartguider via SIMATIC PDM


Bemrk:
Kvikstartguidens indstillinger er indbyrdes forbundet og ndres frst, nr der klikkes p Apply and Transfer (Indfr og overfr) efter det sidste trin. Klik p BACK (Tilbage) for at vende tilbage og ndre en indstilling eller CANCEL (Annuller) for at forlade Kvikstart. bn menuen Device (Apparat) i SIMATIC PDM og vlg den Kvikstart, der passer til applikationen (Niveau, Volumen, Volumen-Linearisering, Flow).

Dansk

Kvikstarts

Betjening via AMS Device Manager


AMS Device Manager er en softwarepakke, der er beregnet til at idriftstte og vedligeholde SITRANS LUT400 og andre procesapparater. Yderligere information kan findes p: http://www.emersonprocess.com/AMS/.

Electronic Device Description (EDD - Elektronisk apparatbeskrivelse)


SITRANS LUT400 krver en EDD for AMS Device Manager version 10.5.

Konfiguration af et nyt apparat via AMS Device Manager


1) EDD kan findes i Apparatkataloget (Device Catalog) under Sensors/Level/Echo/Siemens AG/SITRANS LUT400 (Flere/Niveau/Ekko/Siemens AG/SITRANS LUT400). Tjek produktsiden p vort website p: www.siemens.com/sitransLUT400, under Support>Software Downloads, for at kontrollere, at De har den seneste version af EDD for AMS Device Manager. Gem filerne p computeren og pak den zippede fil ud til en lettilgngelig placering. Start AMS Device Manager Add Device Type (Tilfj apparattype), g til den udpakkede EDD-fil og vlg den. (En brugsanvisning i opstningen af HART-apparater med AMS Device Manager kan downloades fra produktsiden p vort website under Support). I Device Connection View, (Visning af apparattilslutning), hjreklik p apparatets ikon og vlg Scan Device (Scan apparat) for at uploade parametrene fra apparatet. Dobbeltklik p apparatets ikon for at bne opstartsskrmen. Opstartsskrmen viser apparatets identifikationsoplysninger samt et navigationsvindue til venstre p skrmen. Konfigurer SITRANS LUT400 ved hjlp af den passende Kvikstartguide.

2) 3)

4) 5) 6)

7ML19985XU81

SITRANS LUT400 (HART)

Side DA-33

Betjening via FC375/475 (HART)


FC375/FC475 HART Communicator er et brugervenligt hndholdt kommunikationsapparat, der universelt understtter andre HART-apparater, ssom SITRANS LUT400. Menustrukturen af FC375/475 minder meget om strukturen af AMS Device Manager.

Electronic Device Description (EDD - Elektronisk apparatbeskrivelse)


SITRANS LUT400 krver EDD for FC375/FC475.

Konfiguration af et nyt apparat via FC375/FC475.


1) Tjek produktsiden p vort website p www.siemens.com/sitransLUT400, under Support>Software Downloads, for at kontrollere, at De har den seneste version af EDD for FC375/ FC475. Gem filerne p computeren og pak den zippede fil ud til en lettilgngelig placering. For at installere EDD, kr Emersons Field Communicator Easy Upgrade Utility. Konfigurer SITRANS LUT400 ved hjlp af Kvikstartguiden, der passer til applikationen.

Dansk

2) 3) 4)

Betjening via FDT (Field Device Tool)


FDT er en standard, som anvendes i adskillige softwarepakker, der er beregnet til at idriftstte og vedligeholde apparater ssom SITRANS LUT400. To kommercielt tilgngelige FDTs er PACTware og Fieldcare. FDT minder meget om PDM. For at konfigurere et feltapparat via FDT krves apparatets DTM (Device Type Manager - Apparattypemanager). Siemens instrumenter bruger SITRANS DTM og en instrument-EDD, der er skrevet til SITRANS DTM. 1) Installer frst SITRANS DTM p Deres system. Det kan downloades fra: http://www.siemens.com/sitransdtm. Klik p Support og g dernst til Software downloads. Installer SITRANS LUT400 (HART) EDD for SITRANS DTM. Den kan downloades fra produktsiden p vort website: G til www.siemens.com/sitransLUT400 under Support og klik p Software Downloads.

2)

Konfiguration af et nyt apparat via FDT


En brugsanvisning kan downloades fra produktsiden p vort website under Support.

Service og vedligeholdelse
SITRANS LUT400 krver ingen vedligeholdelse eller rengring under normale driftsbetingelser.

Reparation af apparatet og ansvarsbegrnsning


For yderligere oplysninger henvises til indersiden af bagomslaget.

Side DA-34

SITRANS LUT400 (HART)

7ML19985XU81

Elektrisk installation i risikoomrder


Der findes flgende muligheder for elinstallation i risikoomrder: Stvantndelsesbeskyttet elinstallation (Internationalt/Europa): Ikke-antndingsfarlig elinstallation (Canada) Tjek i alle tilflde godkendelsesniveauet p instrumentets skilt.

1. Stvantndelsesbeskyttet elinstallation (Internationalt/Europa):


SITRANS LUT420 7MLxxxx-xxxxx-xxxx Serial No.: GYZ / B1034567 0518 KCC-REM-S49 Power Rating: 100 230V 15% SITRANSLUT400 50/60 Hz, 36 VA (10 W) Contact Rating: 1A/5A @ 250V , Non-Inductive Operating Temperature: 20C to 50C Enclosure: IP65 / TYPE 4X / NEMA 4X II 3D Ex tc IIIC T70C Dc Ex tc IIIC T70C Dc SIRA 12ATEX9143X IECEx SIR 12.0057X Warning: Possible static hazard. Do not rub or clean on site with a dry cloth. Warning: Do not open when energized or when an explosive dust atmosphere is present. Siemens Milltronics Process Instruments, Peterborough Assembled in Canada with domestic and imported parts SITRANS LUT430 7MLxxxx-xxxxx-xxxx Serial No.: GYZ / B1034567 0518 KCC-REM-S49 Power Rating: 100 230V 15% SITRANSLUT400 50/60 Hz, 36 VA (10 W) Contact Rating: 1A/5A @ 250V , Non-Inductive Operating Temperature: 20C to 50C Enclosure: IP65 / TYPE 4X / NEMA 4X II 3D Ex tc IIIC T70C Dc Ex tc IIIC T70C Dc SIRA 12ATEX9143X IECEx SIR 12.0057X Warning: Possible static hazard. Do not rub or clean on site with a dry cloth. Warning: Do not open when energized or when an explosive dust atmosphere is present. Siemens Milltronics Process Instruments, Peterborough Assembled in Canada with domestic and imported parts SITRANS LUT440 7MLxxxx-xxxxx-xxxx Serial No.: GYZ / B1034567 0518 KCC-REM-S49 Power Rating: 100 230V 15% SITRANSLUT400 50/60 Hz, 36 VA (10 W) Contact Rating: 1A/5A @ 250V , Non-Inductive Operating Temperature: 20C to 50C Enclosure: IP65 / TYPE 4X / NEMA 4X II 3D Ex tc IIIC T70C Dc Ex tc IIIC T70C Dc SIRA 12ATEX9143X IECEx SIR 12.0057X Warning: Possible static hazard. Do not rub or clean on site with a dry cloth. Warning: Do not open when energized or when an explosive dust atmosphere is present. Siemens Milltronics Process Instruments, Peterborough Assembled in Canada with domestic and imported parts

Dansk

SITRANS LUT420 7MLxxxx-xxxxx-xxxx Serial No.: GYZ / B1034567 0518 KCC-REM-S49 Power Rating: 10 32V , 10W SITRANSLUT400 Contact Rating: 1A/5A @ 250V , Non-Inductive Operating Temperature: 20C to 50C Enclosure: IP65 / TYPE 4X / NEMA 4X

SITRANS LUT430 7MLxxxx-xxxxx-xxxx Serial No.: GYZ / B1034567 0518 KCC-REM-S49 Power Rating: 10 32V , 10W SITRANSLUT400 Contact Rating: 1A/5A @ 250V , Non-Inductive Operating Temperature: 20C to 50C Enclosure: IP65 / TYPE 4X / NEMA 4X

SITRANS LUT440 7MLxxxx-xxxxx-xxxx Serial No.: GYZ / B1034567 0518 KCC-REM-S49 Power Rating: 10 32V , 10W SITRANSLUT400 Contact Rating: 1A/5A @ 250V , Non-Inductive Operating Temperature: 20C to 50C Enclosure: IP65 / TYPE 4X / NEMA 4X

II 3D Ex tc IIIC T70C Dc Ex tc IIIC T70C Dc SIRA 12ATEX9143X IECEx SIR 12.0057X Warning: Possible static hazard. Do not rub or clean on site with a dry cloth. Warning: Do not open when energized or when an explosive dust atmosphere is present.

II 3D Ex tc IIIC T70C Dc Ex tc IIIC T70C Dc SIRA 12ATEX9143X IECEx SIR 12.0057X Warning: Possible static hazard. Do not rub or clean on site with a dry cloth. Warning: Do not open when energized or when an explosive dust atmosphere is present.

II 3D Ex tc IIIC T70C Dc Ex tc IIIC T70C Dc SIRA 12ATEX9143X IECEx SIR 12.0057X Warning: Possible static hazard. Do not rub or clean on site with a dry cloth. Warning: Do not open when energized or when an explosive dust atmosphere is present.

Siemens Milltronics Process Instruments, Peterborough Assembled in Canada with domestic and imported parts

Siemens Milltronics Process Instruments, Peterborough Assembled in Canada with domestic and imported parts

Siemens Milltronics Process Instruments, Peterborough Assembled in Canada with domestic and imported parts

ATEX-certifikatet, der er angivet p apparatets skilt, kan downloades fra produktsiden p vort website p www.siemens.com/sitransLUT400. IECEx-certifikatet, der er angivet p apparatets skilt, kan lses p IECEx-websitet. G til: http://iecex.iec.ch og klik p Ex Equipment Certificates of Conformity og indtast certifikatnummeret IECEx SIR 12.0057X. Mht. kravene til elinstallationen skal de lokale bestemmelser overholdes. Se ogs ovenstende ATEX-certifikat.

7ML19985XU81

SITRANS LUT400 (HART)

Side DA-35

2. Ikke-antndingsfarlig elinstallation (Canada)


SITRANS LUT420 7MLxxxx-xxxxx-xxxx Serial No.: GYZ / B1034567 Power Rating: 100 230V 15% 50/60 Hz, 36 VA (10 W) Contact Rating: 1A/5A @ 250V , Non-Inductive Operating Temperature: 20C to 50C Enclosure: IP65 / TYPE 4X / NEMA 4X SITRANS LUT430 7MLxxxx-xxxxx-xxxx Serial No.: GYZ / B1034567 Power Rating: 100 230V 15% 50/60 Hz, 36 VA (10 W) Contact Rating: 1A/5A @ 250V , Non-Inductive Operating Temperature: 20C to 50C Enclosure: IP65 / TYPE 4X / NEMA 4X SITRANS LUT440 7MLxxxx-xxxxx-xxxx Serial No.: GYZ / B1034567 Power Rating: 100 230V 15% 50/60 Hz, 36 VA (10 W) Contact Rating: 1A/5A @ 250V , Non-Inductive Operating Temperature: 20C to 50C Enclosure: IP65 / TYPE 4X / NEMA 4X

159134

Class I, Div.2, Gr. A, B, C & D T3C Class II, Div. 2, Gr. F & G Class III Per Drawing A5E03936871

159134

Class I, Div.2, Gr. A, B, C & D T3C Class II, Div. 2, Gr. F & G Class III Per Drawing A5E03936871

159134

Class I, Div.2, Gr. A, B, C & D T3C Class II, Div. 2, Gr. F & G Class III Per Drawing A5E03936871

Siemens Milltronics Process Instruments, Peterborough Assembled in Canada with domestic and imported parts

Siemens Milltronics Process Instruments, Peterborough Assembled in Canada with domestic and imported parts

Siemens Milltronics Process Instruments, Peterborough Assembled in Canada with domestic and imported parts

Dansk

SITRANS LUT420 7MLxxxx-xxxxx-xxxx Serial No.: GYZ / B1034567 Power Rating: 10 32V , 10W Contact Rating: 1A/5A @ 250V , Non-Inductive Operating Temperature: 20C to 50C Enclosure: IP65 / TYPE 4X / NEMA 4X

SITRANS LUT430 7MLxxxx-xxxxx-xxxx Serial No.: GYZ / B1034567 Power Rating: 10 32V , 10W Contact Rating: 1A/5A @ 250V , Non-Inductive Operating Temperature: 20C to 50C Enclosure: IP65 / TYPE 4X / NEMA 4X

SITRANS LUT440 7MLxxxx-xxxxx-xxxx Serial No.: GYZ / B1034567 Power Rating: 10 32V , 10W Contact Rating: 1A/5A @ 250V , Non-Inductive Operating Temperature: 20C to 50C Enclosure: IP65 / TYPE 4X / NEMA 4X

159134

Class I, Div.2, Gr. A, B, C & D T3C Class II, Div. 2, Gr. F & G Class III Per DWG. A5E03936871

159134

Class I, Div.2, Gr. A, B, C & D T3C Class II, Div. 2, Gr. F & G Class III Per DWG. A5E03936871

159134

Class I, Div.2, Gr. A, B, C & D T3C Class II, Div. 2, Gr. F & G Class III Per DWG. A5E03936871

Siemens Milltronics Process Instruments, Peterborough Assembled in Canada with domestic and imported parts

Siemens Milltronics Process Instruments, Peterborough Assembled in Canada with domestic and imported parts

Siemens Milltronics Process Instruments, Peterborough Assembled in Canada with domestic and imported parts

CSA Klasse I, Div 2 tilslutningsdiagram nummer A5E03936871 kan downloades fra produktsiden p vort website p www.siemens.com/sitransLUT400.

Srlige anvisninger vedrrende installation i risikoomrder (Reference: det europiske ATEX Direktiv 94/9/EF, Bilag II, 1/0/6)
Flgende anvisninger glder for udstyr, der er dkket af certifikat nummer SIRA 12ATEX9143X: 1) 2) 3) 4) Hvad angr brug og montage, henvises der til de generelle anvisninger. Dette udstyr er godkendt til brug som Kategori 3D udstyr. Udstyret kan anvendes i risikoomrder 22. Dette udstyr har en maksimal overfladetemperatur p T70 C (ved omgivende +50 C). Der henvises til de gldende normer med hensyn til valget af dette udstyr i forhold til specifikke stvantndelsestemperaturer. 5) Dette udstyr er godkendt til brug ved omgivende temperaturer i intervallet fra -20 til +50 C. 6) Alle kabel- eller kabelrrsindfringer skal overholde kravene i det europiske direktiv 94/9/EF for Gruppe II, Kategori 3D og bevare husets samlede IP-klassificering. 7) Kablet skal vlges i henhold til de gldende normer og skal vre sledes, at dets isolation kan modst husets maksimale overfladetemperatur (T70 C). 8) Dette udstyr betragtes ikke som en sikkerhedsanordning (iflge direktiv 94/9/EF Bilag II, paragraf 1.5). 9) Installationen og inspektionen af dette udstyr skal udfres af tilstrkkeligt uddannet personale i henhold til de gldende normer (EN 60079-14 og EN 60079-17 i Europa). 10) Reparation af dette udstyr skal udfres af tilstrkkeligt uddannet personale i henhold til de gldende normer (fx EN 60079-19 i Europa).

Side DA-36

SITRANS LUT400 (HART)

7ML19985XU81

11) Komponenter, der skal indbygges eller bruges som erstatningskomponenter i udstyret, skal monteres af tilstrkkeligt uddannet personale i henhold til fabrikantens dokumentation. 12) En manuel omgelse kan opns ved at bruge den afbryder, der indgr i bygningens elinstallation.

SPECIELLE BETINGELSER FOR SIKKER BRUG


Bogstavet X i slutningen af certifikatnummeret henviser til flgende srlig(e) betingelse(r) for sikker brug: Alle kabel- eller kabelrrsindfringer skal overholde kravene i det europiske direktiv 94/9/EF for Gruppe II, Kategori 3D og skal bevare husets samlede IP-klassificering. Kablet skal vlges i henhold til de gldende normer og skal vre sledes, at dets isolation kan modst husets maksimale overfladetemperatur (T70 C). Huset er ikke-ledende og kan, under visse ekstreme forhold, ophobe en s stor elektrostatisk ladning, at den kan forrsage antndelse. Brugeren skal srge for, at udstyret ikke installeres p et sted, hvor det kan blive udsat for ydre forhold (ssom hjtryksdamp), der kan medfre en akkumulering af elektrostatiske ladninger p ikkeledende overflader. Udstyret br desuden kun gres rent med en fugtig klud. Udstyret skal enten installeres i et omrde med lav risiko for mekanisk beskadigelse eller, hvis dette ikke er muligt, beskyttes med ekstra afskrmning, der yder passende mekanisk beskyttelse i forhold til applikationen. Alle kontrolenheder i LUT400-serien (LUT420, LUT430, LUT440), der anvendes med fjernmonteret dksel, skal installeres i en indkapsling af passende kategori, der overholder kravene i det europiske direktiv 94/9/EF for Gruppe II, Kategori 3D og niveau IP65 eller derover.

Dansk

7ML19985XU81

SITRANS LUT400 (HART)

Side DA-37

Notater

Dansk
Side DA-38

SITRANS LUT400 (HART)

7ML19985XU81

SITRANS LUT400 (HART) Kurzanleitung


Diese Betriebsanleitung ist eine Kurzfassung der wesentlichen Merkmale und Funktionen des SITRANS LUT400 (HART 1)). Wir empfehlen, die ausfhrliche Version der Betriebsanleitung zu erwerben, damit Sie Ihr Gert optimal nutzen knnen. Die ausfhrliche Betriebsanleitung des SITRANS LUT400 (HART), sowie das Handbuch Kommunikation fr SITRANS LUT400 (HART) steht auf der Produktseite unserer Website zum Download zur Verfgung: www.siemens.com/sitransLUT400. Ein gebundenes Exemplar erhalten Sie bei Ihrer zustndigen Siemens Niederlassung. Wenn Sie Fragen haben, wenden Sie sich bitte an: Siemens AG Siemens Milltronics Process Instruments 1954 Technology Drive, P .O. Box 4225 Peterborough, Ontario, Kanada, K9J 7B1 E-Mail: techpubs.smpi@siemens.com

Copyright Siemens AG 2012. Alle Rechte vorbehalten


Wir fordern Benutzer auf, genehmigte, gebundene Betriebsanleitungen zu erwerben oder die von Siemens Milltronics Process Instruments entworfenen und genehmigten elektronischen Ausfhrungen zu lesen. Siemens Milltronics Process Instruments ist fr den Inhalt auszugsweiser oder vollstndiger Wiedergaben gebundener oder elektronischer Ausfhrungen nicht verantwortlich.

Haftungsausschluss
Wir haben den Inhalt der Druckschrift auf bereinstimmung mit dem beschriebenen Gert geprft. Dennoch knnen Abweichungen nicht ausgeschlossen werden, so dass wir fr die vollstndige bereinstimmung keine Gewhr bernehmen. Die Angaben in dieser Druckschrift werden regelmig berprft, notwendige Korrekturen sind in den nachfolgenden Auflagen enthalten. Fr Verbesserungsvorschlge sind wir dankbar. Technische nderungen vorbehalten.

Deutsch

MILLTRONICS ist ein eingetragenes Warenzeichen von Siemens Milltronics Process Instruments.

Technischer Support
Ein Support steht rund um die Uhr zur Verfgung. Um die Adresse, Telefon- und Faxnummer Ihrer rtlichen Niederlassung von Siemens Automation zu finden, siehe: www.siemens.com/automation/partner: Whlen Sie Deutsch als Sprache und klicken Sie auf Kontakt, um Ihre Produktgruppe zu finden (zum Beispiel: Automatisierungstechnik). Klicken Sie erneut auf die Produktgruppen-Auswahlliste, um auf die Produktfamilie zuzugreifen. Fr Fllstand: +Automatisierungstechnik > +Sensoriksysteme >+Prozessinstrumentierung > +Fllstandmessung > +Kontinuierliche Verfahren. Whlen Sie Ultraschall. Fr Durchfluss: +Automatisierungstechnik > +Sensoriksysteme >+Prozessinstrumentierung > +Durchflussmessung. Whlen Sie Ultraschall. Whlen Sie nun das Land aus, gefolgt von der Stadt/Region. Whlen Sie Technical Support unter Service. Fr technischen Support online, siehe: www.siemens.com/automation/support-request Whlen Sie Deutsch als Sprache und geben Sie das Produkt (SITRANS LUT400) oder die Bestellnummer ein. Klicken Sie auf Suchen und treffen Sie die entsprechende Produktauswahl. Klicken Sie auf Weiter. Geben Sie ein Stichwort zur Beschreibung Ihres Problems ein. Suchen Sie entweder in den zugehrigen Dokumenten oder klicken Sie auf Weiter, um eine Beschreibung Ihres Problems an das Siemens Technical Support Team zu senden. Siemens IA/DT Technical Support Center: Tel. +49 (0)911 895 7222
1)

HARTist ein eingetragenes Warenzeichen von HART Communication Foundation.

7ML19985XU81

SITRANS LUT400 (HART)

Seite DE-1

Sicherheitstechnische Hinweise
Warnhinweise mssen zu Ihrer persnlichen Sicherheit sowie der Sicherheit Dritter und zur Vermeidung von Sachschden beachtet werden. Zu jedem Warnhinweis wird der jeweilige Gefhrdungsgrad angegeben. Das Symbol WARNUNG bezieht sich auf ein Warnsymbol auf dem Produkt und bedeutet, dass bei Nicht-Einhalt der entsprechenden Vorsichtsmanahmen Tod, schwere Krperverletzung und/oder erheblicher Sachschaden eintreten knnen. Dieses Symbol WARNUNG wird verwendet, wenn sich kein entsprechendes Vorsichtssymbol auf dem Produkt befindet, und bedeutet, dass bei Nicht-Einhalt der entsprechenden Vorsichtsmanahmen Tod, schwere Krperverletzung und/ oder erheblicher Sachschaden eintreten knnen. Hinweis: steht fr eine wichtige Information ber das Produkt selbst oder den Teil der Betriebsanleitung, auf den besonders aufmerksam gemacht werden soll.

FCC-Konformitt
Nur fr Installationen in den USA: Richtlinien der FCC (Federal Communications Commission)
WARNUNG: nderungen oder Modifizierungen, die nicht ausdrcklich von Siemens Milltronics genehmigt wurden, knnen zum Entzug der Betriebsgenehmigung des Benutzers fhren. Hinweise: Dieses Gert wurde getestet und mit den fr ein Digitalgert der Class A geltenden Grenzwerten, gem Teil 15 der FCC-Bestimmungen, fr konform erklrt. Diese Grenzwerte sollen einen angemessenen Schutz vor Strungen gewhrleisten, wenn das Gert in einem gewerblichen Umfeld betrieben wird. Das Gert erzeugt und verwendet Funkfrequenzen und kann sie ausstrahlen. Wenn es nicht gem der Betriebsanleitung installiert und betrieben wird, knnen Funkstrungen auftreten. Der Betrieb des Gertes in einem Wohnbereich kann Strungen verursachen. In diesem Fall ist der Benutzer angehalten, die Strung auf eigene Kosten zu beheben.

Deutsch

Seite DE-2

SITRANS LUT400 (HART)

7ML19985XU81

SITRANS LUT400
WARNUNG: Betriebssicherheit und Schutz des SITRANS LUT400 sind nur gewhrleistet, wenn das Gert entsprechend dieser Betriebsanleitung betrieben wird. Hinweis: Dieses Produkt ist fr den Einsatz im Industriebereich vorgesehen. Bei Verwendung in Wohngebieten kann es zu Strungen von verschiedenen Funkanwendungen kommen. SITRANS LUT400 ist ein kompaktes, einkanaliges Ultraschall-Auswertegert fr die kontinuierliche Fllstandmessung von Flssigkeiten, Schlmmen und Schttgtern in groen Messbereichen, sowie die hochgenaue Durchflussberwachung in offenen Gerinnen. Das Gert ist mit allen Ultraschall-Sensoren der Baureihe EchoMaxkompatibel. Es arbeitet in einem Messbereich von 0,3 bis 60 Metern (je nach Sensor). SITRANS LUT400 besitzt ein Local User Interface (LUI, lokale Benutzeroberflche) mit mengefhrter Programmierung. Es zeichnet sich durch eine intuitive, webbasierte Schnittstelle, sowie verschiedene Assistenten fr eine einfache Inbetriebnahme aus. SITRANS LUT400 untersttzt das HART-Kommunikationsprotokoll und verschiedene Softwarepakete fr die Kommunikation ber Fernzugriff [SIMATIC PDM, AMS Device Manager, Field Communicator 375/475 und Field Device Tools (FDTs)]. Das Gert integriert auerdem die neue Generation der Sonic Intelligence. Die beispiellose Messleistung unserer Gerte und die Bedienerfreundlichkeit werden dadurch weiter verbessert.

Technische Daten
Eine vollstndige Liste finden Sie in der Betriebsanleitung des SITRANS LUT400 (HART).

Hilfsenergie
AC-Ausfhrung
AC 100-230 V 15%, 50 / 60 Hz, 36 VA (10W)1) DC 10-32 V, 10W 1 Sicherung: 5 x 20 mm, trge, 0,25A, 250V Sicherung: 5 x 20 mm, trge, 1,6A, 125V
Die angegebene Leistungsaufnahme entspricht einem Maximalwert.

DC-Ausfhrung

Deutsch

1)

Umgebungsbedingungen
Hhe
2000 m max.

Umgebungstemperatur
-20 +50 C (-4 +122 F)

Installationskategorie
II

Verschmutzungsgrad
4

Relative Luftfeuchtigkeit
Fr Auenmontage geeignet (nur mit Gehuse IP65 / Type 4X / NEMA 4X)

Zulassungen
Hinweise: Das Typenschild des Gertes gibt die fr Ihr Gert gltigen Zulassungen an. Verwenden Sie geeignete Conduit- und Kabelverschraubungen, um die Schutzart IP oder NEMA zu gewhrleisten. Allgemein Ex-Bereiche Nicht zndgefhrlich (Non-incendive) (Kanada) Staubexplosionssicher (International/Europa) CSAUS/C, CE, FM, UL listed, C-TICK CSA Class I, Div. 2, Gruppen A, B, C, D; Class II, Div 2, Gruppen F, G; Class III ATEX II 3D Ex tc IIIC T70 C IP65 IECEx SIR 12.0057X SIRA 12ATEX9143X SITRANS LUT400 (HART) Seite DE-3

7ML19985XU81

Einbau
Hinweise: Die Installation darf nur durch qualifiziertes Personal und unter Beachtung der lokalen, gesetzlichen Bestimmungen durchgefhrt werden. Dieses Produkt ist elektrostatisch empfindlich. Befolgen Sie angemessene Verfahren zur Erdung. Alle Feldanschlsse mssen gegen mind. 250 V isoliert sein. Whrend des Betriebes liegt an den Sensorklemmen eine gefhrliche Spannung an. Um die Sicherheitsanforderungen der IEC 61010-1 zu erfllen, sind GleichstromEingangsklemmen von einer Spannungsquelle zu versorgen, die ber eine galvanische Trennung zwischen Ein- und Ausgang verfgt. Die Relaiskontaktklemmen mssen mit Gerten verwendet werden, die keine zugnglichen, stromfhrenden Teile haben. Die Verkabelung muss gem den VDE-Vorschriften erfolgen. Die maximal zulssige Betriebsspannung zwischen benachbarten Relaiskontakten betrgt 250 V. Das Gehuse ist schutzisoliert und besitzt keine Erdverbindung zu der Klemmleiste. Verwenden Sie geeignete Durchfhrungen.

Montageanweisungen
Deutsch
Montageanweisungen weichen je nach Ausfhrung ab (Wand-, Rohr-, DIN-Hutschienenmontage und Gerte mit abgesetzt montiertem Display). Bitte beziehen Sie sich auf die spezifischen Anweisungen fr Ihr Gert.

Hinweis: Bei einer Kabelverlegung im Rohr (Conduit) beachten Sie bitte die Anweisungen zur Kabelverlegung auf Seite 9, bevor Sie den SITRANS LUT400 montieren.

Feldgehuse oder Schalttafeleinbau


Im Lieferumfang aller Konfigurationen des SITRANS LUT400 ist eine Montageplatte enthalten. Die Ausfhrung zum Schalttafeleinbau ist sowohl mit LUI-Anzeige oder Blinddeckel verfgbar.
Montageplatte

Gehuse mit Display-Modul Clip (x 2) Clip (x 2)

Hinweis: Befestigungsteile fr die Wandmontage sind nicht enthalten.

Seite DE-4

SITRANS LUT400 (HART)

7ML19985XU81

Montageplatte

Gehuse mit Blinddeckel Clip (x 2) Clip (x 2)

Hinweis: Befestigungsteile fr die Wandmontage sind nicht enthalten. Gehuserckseite

Langlcher

Deutsch

Wandmontage des Gehuses


1) 2) 3) Lcher fr die vier Schrauben (kundenseitig) markieren und bohren. Mit einem Schraubendreher befestigen. Gleichen Sie die Langlcher auf der Rckseite des Gerts mit den Clips auf der Montageplatte ab. Drcken Sie den LUT400 bndig gegen die Montageplatte und schieben Sie ihn nach unten, bis er fest sitzt.

Schraubenlcher auf der Rckplatte

Bitte beachten Sie:


Empfohlene Gre fr die Befestigung: M8oder 5/16"-Schraube mit Scheibe, Auendurchmesser maximal 17 mm oder 5/8" Empfohlene Montage: direkte Wandmontage. Bei einer anderen Montageflche achten Sie darauf, dass diese das vierfache Gewicht des Gerts tragen knnen MUSS.

7ML19985XU81

SITRANS LUT400 (HART)

Seite DE-5

Abgesetzt montiertes Display


abgesetzter Deckel Dichtung mit Display

4 Befestigungsteile fr den Schalttafeleinbau mitgeliefert

Verbindungskabel SITRANS LUT400

Deutsch

Montage des abgesetzt montierten Displays


Hinweis: Das abgesetzt montierte Display kann bis zu 5 m vom Gert entfernt eingebaut werden. Verwenden Sie dazu zwei optionale Verlngerungskabel (jeweils 2,5 m lang). Siehe ausfhrliche Betriebsanleitung. 1) Verwenden Sie das mitgelieferte Muster und schneiden Sie das erforderliche Loch fr den abgesetzten Deckel mit LUI-Display aus. Legen Sie die Formdichtung in den Deckel und passen Sie sie an die Montagelcher an. Richten Sie die Rckseite des abgesetzten Deckels bndig mit dem Schalttafelausschnitt aus. Lcher fr die vier Schrauben (mitgeliefert) markieren und bohren. Mit einem Schraubendreher und Schlssel befestigen. Hinweis: Empfohlenes Drehmoment an Befestigungsschrauben fr gute Abdichtung: 1,1 N/m 10 in-lbs

2)

Bitte beachten Sie:


Empfohlene Montage: Montage an Schalttafel, bis zu 5 m Abstand vom Gert. Bei einer anderen Montageflche achten Sie darauf, dass diese das vierfache Gewicht des Gerts tragen knnen MUSS. Hinweis: Befestigungsteile im Lieferumfang: M5-Schraube, Dichtscheibe, M5-Beilagscheibe und Mutter. Diese Befestigungsteile sind erforderlich, um die Schutzart IP65 des abgesetzt montierten Deckels zu erhalten. Seite DE-6 SITRANS LUT400 (HART) 7ML19985XU81

Rohrmontage

Montageplatte U-Bgel Rohrschelle Gehuse mit Display-Modul

Rohr Hinweis: Befestigungsteile fr die Rohrmontage sind nicht enthalten.

Gehuseeinbau
1) 2) Befestigen Sie die Montageplatte am Rohr mit dem U-Bgel einer Rohrschelle (kundenseitig) passend zur Rohrweite. Die Schrauben mit einem Schlssel befestigen. Nicht zu stark anziehen, um die Platte nicht zu verkrmmen oder zu verbiegen. Dies knnte die Befestigung des LUT400 an der Montageplatte erschweren. Befestigen Sie das Gert an der Montageplatte (siehe Beschreibung in Schritt 3 unter Wandmontage des Gehuses auf Seite 5).

Deutsch

3)

Bitte beachten Sie:


Empfohlene Montage: direkt am horizontalen oder vertikalen Rohr. Bei einer anderen Montageflche achten Sie darauf, dass diese das vierfache Gewicht des Gerts tragen knnen MUSS. Empfohlene Rohrmae: Maximal: 3" Rohr, minimal: 3/4" Rohr Empfohlene Gre fr die Befestigung: U-Bgel: Maximal: 3" Rohrgre mit M8- oder 3/8"-Gewinde Minimal: 3/4" Rohrgre mit M6- oder 1/4"-Gewinde Sechskantmuttern: M6 oder 1/4" bis M8 oder 3/8" Scheibe: Maximal: 16 mm oder 13/16" Auendurchmesser.

7ML19985XU81

SITRANS LUT400 (HART)

Seite DE-7

DIN-Hutschienenmontage
Gehuserckseite

DIN-Hutschiene

Clips (x 2)

Schlitz

Fhrung der DIN-Hutschiene

Deutsch

Gehuseeinbau
1) 2) 3) Das Gehuseoberteil zur Hutschiene hin anwinkeln und etwas oberhalb der Schienen- oberseite platzieren. Das Gehuse nach unten gegen die Hutschiene schieben, um die Clips auf der Gehuserckseite an der Hutschiene oben festzuhaken. Drcken Sie das Gert bndig gegen die Hutschiene, bis die Fhrung der Hutschiene einrastet. Das Gehuse wird dadurch sicher an der Hutschiene befestigt.

Bitte beachten Sie:


Empfohlene Montage: direkt an der horizontalen Hutschiene. Empfohlene Mae der DIN-Hutschiene: TH 35-7,5 oder TH 35-15 nach IEC 60715. Die Hutschiene MUSS das viermalige Gewicht des SITRANS LUT400 tragen knnen.

Gehuse ausbauen
1) Bringen Sie von der Gertevorderseite aus einen Schraubendreher im Schlitz unterhalb der Hutschienenfhrung an und stemmen Sie ihn nach unten. Dadurch wird die Fhrung vom Unterteil der Hutschiene gelst. Halten Sie die Fhrung nach unten und drcken Sie gleichzeitig das Gehuse nach oben, um die Clips oben aus der Hutschiene zu befreien.

2)

Seite DE-8

SITRANS LUT400 (HART)

7ML19985XU81

Vorbereitung fr die Kabeleinfhrung


Kabel knnen ber Schutzrohr (Conduit) verlegt oder ber Kabelverschraubungen in das Gehuse eingefhrt werden. Um die Kabeleinfhrungen zu ffnen, folgen Sie bitte den Schritten 1 bis 5. Fahren Sie dann fort bei Schritt 6 ff. 1) Stellen Sie sicher, dass der Gehusedeckel geschlossen und die Schrauben angezogen sind. 2) Die Spitze des Schraubendrehers in die Nut am Auendurchmesser der Ausschlagffnung (siehe Abbildung weiter unten) setzen. 3) Mit der Handflche auf das Ende des Schraubendrehers schlagen, um die Kabeleinfhrung auszubrechen. 4) Lsen Sie die Schrauben und nehmen Sie den Gehusedeckel ab. 5) Entfernen Sie mgliche Kunststoffteile im Gehuse. Achten Sie darauf, dass die Elektronik nicht durch elektrostatische Aufladung oder beim Ausbrechen der Kabeleinfhrungen beschdigt wird.
SITRANS LUT400

Kabeleinfhrungen

Ausschlagffnung (x 3)

Nut fr Schraubendreher (x 3)

Hinweise: Nach Entfernen der Ausschlagffnungen betrgt der Durchmesser der Kabeleinfhrung 21,4 mm bis 21,6 mm. M20-Kabelverschraubungen (Durchmesser 20 mm) und 1/2" NPTKabeleinfhrungen (Durchmesser 21,3 mm) passen in diese ffnung. Whlen Sie die geeignete Abdichtung fr die ffnungen mit Sorgfalt aus. Empfohlen wird eine Flachdichtung (anstelle eines O-Rings). Bei Verwendung anderer Kabelverschraubungen liegt es in der Verantwortung des Kunden, die Schutzart IP65 an den Kabeleinfhrungen zu gewhrleisten.

Deutsch

Bei einer Kabelverlegung im Schutzrohr (Conduit)


(Fortsetzung von Schritt 1 bis 5 oben) 6) Nach der Vorbereitung fr die Kabel- einfhrung (s. Schritte 1 bis 5 oben) verwenden Sie fr die Kabeleinfhrung ins Gehuse bei wassergeschtzten Applikationen nur zugelassene Kabelverschraubungen geeigneter Gre. (Rohrgre: 1/2" NPT.) 7) Bringen Sie den Gehusedeckel wieder an und ziehen Sie die Schrauben fest.

Bei offener Kabelverlegung und Verwendung von Kabelverschraubungen


(Fortsetzung von Schritt 1 bis 5 oben) 6) Nach der Vorbereitung fr die Kabeleinfhrung (s. Schritte 1 bis 5 oben) ffnen Sie die Kabelverschraubungen und befestigen Sie sie gut am Gehuse. 7) Kabel durch die Verschraubungen ziehen. Halten Sie die Kabel fr Hilfsenergie und Signalbertragung getrennt und schlieen dann die Drhte an die Klemmleisten an. 8) Die Verschraubungen so anziehen, dass sie gut abdichten. 9) Bringen Sie den Gehusedeckel wieder an und ziehen Sie die Schrauben fest. Seite DE-9

7ML19985XU81

SITRANS LUT400 (HART)

Anschluss
Sicherheitshinweise zum Anschluss
Detaillierte Anschlusshinweise finden Sie in der ausfhrlichen Betriebsanleitung. Alle Systembestandteile mssen gem den Anweisungen installiert werden. Schlieen Sie alle Kabelabschirmungen an die Schirmanschlussklemmen des LUT400 an gekennzeichnet). Zur Vermeidung von Potentialverschlep(am Gert mit dem Symbol pungen mssen die Kabelabschirmungen richtig geerdet werden. Offenliegende Leiter auf abgeschirmten Kabeln sollten so kurz wie mglich sein, um Gefahr von Einstreuungen und Strgeruschen zu verringern.

Anschlieen des SITRANS LUT400


WARNUNGEN:

berprfen Sie die Zulassungen auf dem Typenschild Ihres Gertes. Verwenden Sie geeignete Conduit- und Kabelverschraubungen, um die Schutzart IP oder NEMA zu gewhrleisten.
Hinweise: Eine getrennte Leitungsverlegung kann erforderlich sein, um Standardanforderungen an den Anschluss oder elektrische Richtlinien zu erfllen. Zugang zum Anschlussraum:

Deutsch

1) 2) 3)

ffnen Sie den Bajonettverschluss durch eine Vierteldrehung. Den Deckel nach links ffnen (Scharniere). Sie knnen den Deckel entweder offen lassen (an den Scharnieren befestigt), oder von den Scharnieren lsen und beiseite legen, um an den Anschlussraum zu gelangen.

4) 5) 6)

Nehmen Sie die Anschlsse entsprechend der folgenden Anweisungen vor. Nach Beenden des Anschlusses setzen Sie den Gehusedeckel wieder auf. Schlieen Sie den Bajonettverschluss.

Seite DE-10

SITRANS LUT400 (HART)

7ML19985XU81

Anschlussraum
ber die Klemmleiste des LUT400 knnen alle Ein- und Ausgnge gleichzeitig angeschlossen werden. Die Klemmleisten knnen herausgenommen werden, um den Anschluss zu erleichtern. VORSICHT: Achten Sie beim erneuten Einbau darauf, die Klemmleisten an der richtigen Stelle abzuschlieen. Andernfalls knnen Schden des Gerts oder externer, angeschlossener Betriebsmittel die Folge sein. Hinweis: Empfohlenes Drehmoment an den Klemmschrauben: 0,56 - 0,79 N m 5 - 7 in-lbs Die Schrauben drfen nicht zu fest angezogen werden. ower
TS-3 TRANSDUCER DISCRETE INPUTS INPUT POWER

Spannungsversorgung
14

8 1

9 2

10

11 0.25 A, 250 V SLOW BLOW

12 L1

13 L2/N 100 - 230 V

DC SYNC

SHLD

WHT

BLK

BLK

WHT

Klemmleisten
mA OUT/HART

Batterie

RELAYS

BATTERY BR2032

DISPLAY 15 16 RLY 1 17 18 19 20 21 22 23 24

RLY 2

RLY 3 ACT

PAS SERVICE

USBAnschluss

Deutsch

Display-Kabel

mA HARTAnschluss

Hilfsenergie
WARNUNGEN: Um die Sicherheitsanforderungen der IEC 61010-1 zu erfllen, sind die Gleichstrom-Eingangsklemmen von einer Spannungsquelle zu versorgen, die ber eine galvanische Trennung zwischen Ein- und Ausgang verfgt. Alle Feldanschlsse mssen entsprechend der angelegten Spannung isoliert sein. Wichtig! Vor dem ersten Anlegen der Stromversorgung an den SITRANS LUT400 muss sichergestellt sein, dass angeschlossene Alarm-/Steuergerte ausgeschaltet sind. Vor deren Inbetriebnahme muss eine einwandfreie Funktion des Messsystems gewhrleistet sein.

7ML19985XU81

SITRANS LUT400 (HART)

Seite DE-11

SITRANS LUT400 steht als AC- oder DC-Ausfhrung zur Verfgung.

Wechselstrom (AC)
INPUT POWER

Gleichstrom (DC)
INPUT POWER

12

13

14

12 10 - 32 V

13

14

L1 L2/N 100 - 230 V

AC: AC 100-230 V 15%, 50/60 Hz, 36 VA (10W)

DC: DC 10-32 V, 10W

Hinweis: Prfen Sie, dass das Gert sicher geerdet ist. 1) 2) 3) Fr den Spannungsanschluss isolieren Sie den Kabelmantel ca. 70 mm (2.75") vom Ende des Kabels ab und fhren Sie die Adern durch die Kabelverschraubung ein 1). Anschluss der Drhte wie abgebildet: Die Polaritt ist unterhalb der Klemmleiste gekennzeichnet. Nehmen Sie die Erdung des Gerts entsprechend lokaler Richtlinien vor. Hinweise fr AC-Spannungsanschlsse: Die Anlage muss durch eine 16 A Sicherung oder einen Leitungsschutzschalter kundenseitig abgesichert sein. Ein Schalter als Trennvorrichtung fr die Anschlussspannung (mit entsprechender Kennzeichnung) muss in der Nhe des Gertes und fr den Bediener leicht erreichbar angebracht sein.

Deutsch

Die Batterie
Eine Batterie ist im Lieferumfang des SITRANS LUT400 enthalten. (Die Abbildung auf Seite 11 zeigt, wo sich die Batterie befindet.) Die Batterie (BR2032) hat eine voraussichtliche Lebensdauer von 10 Jahren. Sie wird von der Umgebungstemperatur beeinflusst. Bei einem Spannungsverlust des LUT400 hlt die Batterie den Betrieb der Echtzeituhr des Gerts aufrecht, bis die Spannung wieder hergestellt ist. (Angaben zum Austausch der Batterie finden Sie in der Betriebsanleitung.)

1)Bei einer Kabeleinfhrung ber Schutzrohr (Conduit) verwenden Sie nur zugelassene Rohrverschraubun-

gen geeigneter Gre fr wassergeschtzte Applikationen.

Seite DE-12

SITRANS LUT400 (HART)

7ML19985XU81

Ultraschall-Sensoren
Achtung: Whrend des Betriebes liegt an den Sensorklemmen eine gefhrliche Spannung an. Das Sensorkabel muss in einem geerdeten Metallrohr, getrennt von anderen Leitungen (auer dem Kabel des Temperaturfhlers TS-3, falls verwendet) verlegt werden. Hinweise: Aufgrund der hohen Spannung, die auf die Abschirmung des Koaxialkabels bertragen wird, ist mit dem SITRANS LUT400 kein Koaxialkabel zu verwenden. Den Schirm des LUT400 und den weien Draht getrennt, nicht an den gleichen Klemmen anschlieen. Hinweise in Sensoranleitungen bezglich des gemeinsamen Anschlusses von Schirm und weiem Draht treffen hier nicht zu.
TS-3 TRANSDUCER

SHLD

BLK

WHT

Relais
Die Relaiskontakte sind in abgefallenem Zustand abgebildet. Fr alle Relais kann eine positive oder negative Logik eingerichtet werden (siehe Parameter 2.8.11.Relaislogik in der Betriebsanleitung).
RELAYS

BLK

WHT

Deutsch

Spannungsausfall
Relais 2 und 3 sind Schlieer. Relais 1 kann als ffner oder Schlieer angeschlossen werden. Bei einem Verlust der Eingangsleistung kehren die Relais in ihren normalen Zustand zurck.

15

16 RLY 1

17

18

19

20

21

RLY 2

RLY 3

Relaisnennleistungen
ein Wechselkontakt (ffner oder Schlieer) (Relais 1), 1A bei AC 250 V, ohmsche Last, 3A bei DC 30 V zwei Schlieer (Relais 2,3), 5A bei AC 250 V, ohmsche Last, 3A bei DC 30 V

7ML19985XU81

SITRANS LUT400 (HART)

Seite DE-13

Anschluss HART mA Ausgang (HART)


mA OUT/HART

Fr eine AKTIVE HARTVerbindung (ber die interne Spannungsversorgung des LUT400) schlieen Sie die Klemmen 22 und 23 an.

22

23

24

Fr eine PASSIVE HARTVerbindung (ber externe Spannungsversorgung) schlieen Sie die Klemmen 23 und 24 an.

ACT

PAS

Nhere Angaben finden Sie im Abschnitt mA Ausgangsparameter (2.5.Stromausgang) in der Parameterbeschreibung der ausfhrlichen Betriebsanleitung.

Anschluss bei Einbau in Ex-Bereichen


Siehe Seite 35.

Inbetriebnehmen des SITRANS LUT400


Wir empfehlen die Konfiguration Ihrer Applikation in folgender Reihenfolge: Starten Sie zuerst den Schnellstartassistenten fr Ihre Applikation (Fllstand, Volumen, Durchfluss). Stellen Sie dann ggf. die Pumpen ber den Pumpensteuerassistenten ein. Zuletzt sind Alarme oder sonstige Steuerfunktionen, Summierer und Probenehmer zu konfigurieren, mit Bezug auf die jeweiligen Parameter (siehe ausfhrliche Betriebsanleitung). Schnellstartassistenten knnen auf zwei Arten aufgerufen werden: am Gert (siehe Schnellstartassistenten ber LUI auf Seite 17) ber Fernzugriff (siehe Weitere Schnellstartassistenten auf Seite 17)

Deutsch

Einschalten des SITRANS LUT400


Gert einschalten. SITRANS LUT400 startet automatisch im Messmodus. Auf dem Display erscheint zuerst das Siemens-Logo und anschlieend die aktuelle Firmware-Version des LUI. Gleichzeitig wird der erste Messwert verarbeitet. Bei der ersten Gertekonfiguration werden Sie aufgefordert, eine Sprache zu whlen (Englisch, Deutsch, Franzsisch, Spanisch oder Chinesisch). Stellen Sie das korrekte Datum und die Uhrzeit ein, bevor Sie das Gert konfigurieren. (nderungen ber Parameter 2.14.Datum und Uhrzeit. Nhere Angaben in der Betriebsanleitung.)

Seite DE-14

SITRANS LUT400 (HART)

7ML19985XU81

Die Anzeige (LCD) Anzeige im Messmodus: Normalbetrieb 1) Fllstand


1 10
SITRANS LUT400 FLLSTAND

2
1/2

4 5

18.91
1

6
2

5 SPEICHERN VON PARAMETERN

Durchfluss
1 10 11 12
SITRANS LUT400 DURCHFL... BERFAL...
1 2 3

3
1/2
L/S M

4 5 6

500313..
60.000

TAGESSUMM...

999999.99

5 SPEICHERN VON PARAMETERN

1 TAG (Messstellenbezeichnung) 2 Messwert (Fllstand, Leerraum, Abstand, Volumen, Durchfluss oder berfallhhe) 3 Anzeigewert [Primrvariable (PV)=1 von 2, Sekundrvariable (SV)=2 von 2] 4 Symbol Umschalten1) fr PV oder SV 5 Einheiten 6 Balkenanzeige zur Angabe des Fllstands 7 Zusatz-Anzeigefeld fr konfigurierte Relais (links) und diskrete Eingnge (rechts) 8 Textbereich zur Anzeige von Statusmeldungen 9 Symbol Gertezustand 10 Gewhlte Betriebsart (primr): Fllstand, Leerraum, Abstand, Volumen, berfallhhe oder Durchfluss 11 Betriebsart sekundr = berfallhhe bei Betriebsart (primr) = Durchfluss 12 Summiererwerte: abwechselnde Anzeige von Tagessummierer und laufendem Summierer

Deutsch

Anliegen eines Fehlers


0 LOE 8 Textbereich zur Anzeige eines Fehlercodes oder einer Fehlermeldung 9 Anzeige des Symbols Service erforderlich

Anzeige im Programmiermodus Navigationsansicht


Eine sichtbare Menleiste zeigt aktuelles ULTRASCHALL-SENSOR 2.1.6 Parameteran, dass die Menliste zu lang ist, Men/ KEIN SENSOR nummer ParameterXRS-5 um alle Eintrge anzuzeigen. name XPS-10 Die Hhe des Eintragsbalkens auf EintragsXPS-15 aktueller XCT-8 der Menleiste gibt die Lnge der balken XCT-12 Eintrag XPS-30 Menliste an: je hher der Balken, Menleiste desto weniger Eintrge. Die Position des Eintragsbalkens gibt die ungefhre Position des aktuellen Eintrags in der Liste an. Befindet sich der Balken halb unten in der Menleiste, bedeutet dies, dass der aktuelle Eintrag etwa in der Mitte der Liste ist. Je hher der Balken, desto weniger Eintrge.

1)Drcken Sie Taste

oder

zum Umschalten.

7ML19985XU81

SITRANS LUT400 (HART)

Seite DE-15

Parameteransicht
Men, in dem sich der Parameter befindet Parametername
SENSOR
EINHEIT BETRIEBSART BETRIEBSART SEKUNDR FREQUENZ

Bearbeitungsansicht
2.1.6

Parameternummer Parameterwert/ -auswahl

ULTRASCHALL-SENSOR
KEIN SENSOR XPS-10 XPS-15 XCT-8 XCT-12 XPS-30

2.1.6

XRS-5

ULTRASCHALL-SENSOR

XRS-5

PROGRAMMIER-Modus
Anhand der Bedientasten am Gert drcken Sie zum Aufruf des Programmiermodus und ffnen Sie Men-Ebene 1. Zum Blttern durch das Men verwenden Sie , , , .

Eine aufgelistete Option whlen


a) Rufen Sie den gewnschten Parameter auf. b) Drcken Sie den RECHTS-Pfeil , um den Bearbeitungs-Modus zu starten. Die aktuelle Auswahl ist hervorgehoben. c) Blttern Sie auf eine neue Auswahl. d) Drcken Sie den RECHTS-Pfeil zur Besttigung. Das LCD kehrt auf die Parameteransicht zurck und zeigt die neue Auswahl an.

Einen numerischen Wert ndern Deutsch


a) Rufen Sie den gewnschten Parameter auf, zum Beispiel Unterer Kalibrierungspunkt (2.2.1). b) Bei seiner Auswahl wird der aktuelle Wert angezeigt. c) Drcken Sie den RECHTS-Pfeil , um den Bearbeitungsmodus zu starten. Die Cursorposition wird hervorgehoben.
UNTERER KALIB.PKT 2.2.1

ESC

+ _ _ _ _ _ _ 2. 0 M

OK

d) Mit dem LINKSund RECHTS-Pfeil bewegen Sie den Cursor auf die zu ndernde Stelle. e) Sobald eine Stelle hervorgehoben (gewhlt) ist, kann sie mit dem Pfeil nach OBEN oder UNTEN f) jeweils erhht bzw. vermindert werden. und UNTEN Bei gewhlter Dezimalstelle kann diese mit dem Pfeil nach OBEN verschoben werden. g) Zum Ausstieg, ohne Ihre nderungen zu speichern, drcken Sie den LINKS-Pfeil solange, bis ESC hell leuchtet. Drcken Sie erneut den LINKS-Pfeil , um auszusteigen, ohne die nderungen zu speichern. Andernfalls, wenn der neue Parameterwert korrekt ist, drcken Sie den RECHTS-Pfeil wird. solange, bis OK hervorgehoben

h) Drcken Sie den RECHTS-Pfeil zur Besttigung des neuen Werts. Das LCD kehrt auf die Parameteransicht zurck und zeigt die neue Auswahl an. Der Genauigkeit halber nachprfen.

Seite DE-16

SITRANS LUT400 (HART)

7ML19985XU81

Schnellstartassistenten
Um den SITRANS LUT400 fr Fllstand-, Volumen- (Standard-Behlterformen) oder Durchflussapplikationen zu konfigurieren, verwenden Sie die Assistenten, die unter Schnellstartassistenten ber LUI beschrieben sind. Ein Assistent fr Applikationen mit komplexeren Behlterformen ist ber SIMATIC PDM verfgbar. Siehe Schnellstartassistenten ber SIMATIC PDM auf Seite 33.

Weitere Schnellstartassistenten
Andere Schnellstartassistenten, die verschiedene Softwarepakete verwenden, sind ebenfalls verfgbar 1): SIMATIC PDM (HART) AMS (HART) FC375/475 (HART) FDTs (HART) Parameterkonfigurationstabellen (zur Datenaufzeichnung vor der Konfiguration des Gerts mit dem Schnellstartassistenten) sind auf unserer Website verfgbar. Siehe www.siemens.com/sitransLUT400 > Support > Applikationsbeispiele.

Schnellstartassistenten ber LUI


1) 2) Taste zum Aufruf des Programmiermodus. Whlen Sie Assistenten (1.), Quick Start (1.1.), und dann den passenden Schnellstart: QS Fllstand (1.1.1.), QS Volumen (1.1.2.) oder QS Durchfluss (1.1.3.). [Der Schnellstartassistent Durchfluss erscheint nur auf dem LUI der Ausfhrungen LUT430 (Pumpen und Durchfluss) und LUT440 (OCM).] Folgen Sie den Schritten und whlen Sie Fertigstellen, um die genderten Schnellstartparameter zu speichern und in das Men der Programmierung zurckzukehren. Dann drcken Sie dreimal zur Rckkehr in den Messmodus. Hinweise: Die Einstellungen des Schnellstartassistenten sind zusammenhngend und nderungen werden erst wirksam, wenn Sie Fertigstellen im letzten Schritt whlen. Verwenden Sie den Schnellstartassistenten nicht, um einzelne Parameter zu ndern. Beziehen Sie sich dazu auf den Abschnitt Parameter der Betriebsanleitung. (Kundenspezifische Anpassung an Ihre Anwendung erst nach Beendigung des Schnellstarts durchfhren). Nhere Angaben zu den Parametern der Assistenten finden Sie in der Betriebsanleitung.

Deutsch

3)

1. Assistenten 1.1. Quick Start 1.1.1. QS Fllstand


Mit diesem Assistenten werden einfache Fllstandapplikationen konfiguriert. a) Drcken Sie den RECHTS-Pfeil zum Aufruf des PROGRAMMIER-Modus und ffnen Sie die Menebene 1: HAUPTMEN. b) Drcken Sie den RECHTS-Pfeil c) Drcken Sie den RECHTS-Pfeil zweimal, um zum Meneintrag 1.1.1 zu gelangen. , um QS Fllstand zu ffnen.

1)Siehe Handbuch Kommunikation fr SITRANS LUT400 (HART)

7ML19985XU81

SITRANS LUT400 (HART)

Seite DE-17

d) In jedem Schritt knnen Sie mit dem Pfeil Vorgabewerte akzeptieren nach UNTEN und direkt zum nchsten Eintrag gelangen, oder mit dem RECHTS-Pfeil den Bearbeitungsmodus starten: die aktuelle Auswahl ist hervorgehoben. e) Blttern Sie auf den gewnschten Eintrag und drcken Sie den RECHTS-Pfeil , um die nderung zu speichern, gefolgt vom Pfeil nach UNTEN f) , um fortzufahren.

SCHNELLSTART FLLSTAND
ULTRASCHALL-SENSOR XRS-5
ZURCK ABBRECHEN WEITER BEARBEITEN

SCHNELLSTART FLLSTAND
ULTRASCHALL-SENSOR

KEIN SENSOR XPS-10 XPS-15 XCT-8 XCT-12

XRS-5

Sie knnen jederzeit den Pfeil nach OBEN LINKS-Pfeil Start des Assistenten QS Fllstand

drcken, um zurckzufahren, oder den

, um den Assistenten abzubrechen.

Zeigt den Typ des auszufhrenden Assistenten an.

Optionen

ABBRECHEN, START

Ultraschall-Sensor

Angabe des angeschlossenen Siemens Sensortyps.

Optionen Deutsch
Betriebsart

KEIN SENSOR, XRS-5, XPS-10, XPS-15, XCT-8, XCT-12, XPS-30, XPS-40, XLT-30, XLT-60, STH Voreinstellung: KEIN SENSOR

Einstellung der fr die Applikation erforderlichen Messart (und des entsprechenden mA Ausgangs).
Sensor-Bezugspunkt Oberer Kalibrierungspunkt Leerraum

Abstand

Fllstand Unterer Kalibrierungspunkt

Endbereich

Modus

Beschreibung

Bezugspunkt
Unterer Kalibrierungspunkt (Nullpunkt des Prozesses) Oberer Kalibrierungspunkt (Vollpunkt des Prozesses) Sensor-Bezugspunkt (Fortsetzung auf nchster Seite)

FLLSTAND * Hhe des Materials LEERRAUM ABSTAND Abstand zur Materialoberflche

Seite DE-18

SITRANS LUT400 (HART)

7ML19985XU81

(Forts. von vorhergehender Seite)

Modus

Beschreibung
NICHT auswhlen. Wenn als Wert der Betriebsart ANDERE erscheint, ist das Gert als Fllstand-Auswertegert konfiguriert, aber zuvor auf einen anderen Modus als FLLSTAND, LEERRAUM oder ABSTAND eingestellt worden. Um mit dem Schnellstartassistenten Fllstand fortzufahren, muss FLLSTAND, LEERRAUM oder ABSTAND eingestellt sein.

Bezugspunkt

ANDERE

Temperaturmessung

Quelle des angezeigten Temperaturwertes fr die Korrektur der Schallgeschwindigkeit.

Optionen

ULTRASCHALLSENSOR, TEMPERATURVORGABE, EXTERNER TS-3, MITTELWERT VON SENSOREN Voreinstellung: ULTRASCHALL-SENSOR

Temperaturvorgabe

Dieser Parameter wird bentigt, wenn kein Temperaturmessgert eingesetzt wird.

Wert

Bereich: -100,0 ... +150,0 C Voreinstellung: +20,0 C

Deutsch

Dieser Parameter wird nur angezeigt, wenn TEMPERATURVORGABE in Temperaturmessung gewhlt ist. Einheit

Maeinheiten des Sensors.

Optionen

M, CM, MM, FT, IN Voreinstellung: M

Hinweis: Fr dieses Beispiel werden alle Werte in Metern (m) angenommen.


Oberer Kalibrierungspunkt

Abstand vom Sensor-Bezugspunkt zum Oberen Kalibrierungspunkt: entspricht in der Regel dem Vollpunkt des Prozesses.

Wert

Bereich: 0,000 ... 60,000 Voreinstellung: 0,000

Unterer Kalibrierungspunkt

Abstand vom Sensor-Bezugspunkt zum Unteren Kalibrierungspunkt: entspricht in der Regel dem Nullpunkt des Prozesses.

Wert

Bereich: 0,000 ... 60,000 Voreinstellung: 60,000

7ML19985XU81

SITRANS LUT400 (HART)

Seite DE-19

Reaktionszeit

Einstellung der Geschwindigkeit, mit der das Gert auf nderungen im Zielbereich reagiert.

Hinweise:

Die Reaktionszeit kann nur ber den Schnellstartassistenten eingestellt werden. Alle nderungen der Parameter Befllgeschwindigkeit pro Minute (2.3.1.), Entleergeschwindigkeit pro Minute (2.3.2.) oder Dmpfungsfilter (2.3.3.), die nach Beenden des Assistenten vorgenommen werden, ersetzen die Einstellung der Reaktionszeit. Reaktionszeit wird immer in m/min angezeigt.
LANGSAM (0,1 M/MIN) MITTEL (1,0 M/MIN) SCHNELL (10 M/MIN) Voreinstellung: LANGSAM (0,1 M/MIN)

Optionen

Die Einstellung sollte die max. Geschwindigkeit beim Befllen oder Entleeren (es gilt der grere Wert) gerade etwas bersteigen. Niedrigere Werte ergeben eine hhere Genauigkeit, whrend hhere Werte schnellere Fllstandschwankungen bercksichtigen knnen. Ende des Assistenten QS Fllstand

Fr einen erfolgreichen QS mssen alle nderungen angewendet werden.

Optionen Deutsch

ZURCK, ABBRECHEN, FERTIGSTELLEN (Anzeige kehrt auf Men 1.1 Schnellstart zurck, wenn der Schnellstart erfolgreich beendet oder abgebrochen ist. Bei Auswahl von ABBRECHEN werden keine nderungen an das Gert geschrieben.)

Um Schnellstart-Werte an das Gert zu bertragen und in das Men der Programmierung zurckzukehren, drcken Sie den Pfeil nach UNTEN (Fertigstellen). Drcken Sie daraufhin dreimal den LINKS-Pfeil , um in den Messmodus zurckzukehren.

1.1.2. QS Volumen
Mit diesem Assistenten knnen Volumenapplikationen mit Standard-Behlterformen konfiguriert werden. a) Drcken Sie den RECHTS-Pfeil zum Aufruf des PROGRAMMIER-Modus und ffnen Sie die Menebene 1: HAUPTMEN. b) Drcken Sie den RECHTS-Pfeil zweimal, um zum Meneintrag 1.1.1 zu gelangen. c) Drcken Sie den Pfeil nach UNTEN und den RECHTS-Pfeil , um QS Volumen zu ffnen. d) In jedem Schritt knnen Sie mit dem Pfeil nach UNTEN SCHNELLSTART VOLUMEN ULTRASCHALL-SENSOR Vorgabewerte akzeptieren und direkt zum nchsten XRS-5 Eintrag gelangen, oder mit dem RECHTS-Pfeil den Bearbeitungsmodus starten: die aktuelle Auswahl ist hervorgehoben. SCHNELLSTART VOLUMEN e) Blttern Sie auf den gewnschten Eintrag und drcken KEIN SENSOR , um die nderung zu speiSie den RECHTS-Pfeil XRS-5 XPS-10 chern, gefolgt vom Pfeil nach UNTEN , um fortzufahXPS-15 ren. XCT-8
ZURCK ABBRECHEN BEARBEITEN WEITER

ULTRASCHALL-SENSOR

XCT-12

Seite DE-20

SITRANS LUT400 (HART)

7ML19985XU81

f)

Sie knnen jederzeit den Pfeil nach OBEN drcken, um zurckzufahren, oder den LINKS-Pfeil , um den Assistenten abzubrechen. Start des Assistenten QS Volumen

Zeigt den Typ des auszufhrenden Assistenten an.

Optionen

ABBRECHEN, START

Ultraschall-Sensor

Angabe des angeschlossenen Siemens Sensortyps.

Optionen

KEIN SENSOR, XRS-5, XPS-10, XPS-15, XCT-8, XCT-12, XPS-30, XPS-40, XLT-30, XLT-60, STH Voreinstellung: KEIN SENSOR

Temperaturmessung

Quelle des angezeigten Temperaturwertes fr die Korrektur der Schallgeschwindigkeit.

Optionen

ULTRASCHALLSENSOR, TEMPERATURVORGABE, EXTERNER TS-3, MITTELWERT VON SENSOREN Voreinstellung: ULTRASCHALL-SENSOR

Temperaturvorgabe

Dieser Parameter wird bentigt, wenn kein Temperaturmessgert eingesetzt wird.

Wert

Bereich: -100,0 ... +150,0 C Voreinstellung: +20,0 C

Deutsch

Dieser Parameter wird nur angezeigt, wenn TEMPERATURVORGABE in Temperaturmessung gewhlt ist. Behlterform

Definiert die Behlterform und ermglicht eine Volumenberechnung anstelle einer Fllstandberechnung des SITRANS LUT400. Bei der Auswahl OHNE erfolgt keine Volumenberechnung. Whlen Sie die Behlterform, die dem zu messenden Behlter entspricht.

Optionen

OHNE, LINEAR, ZYLINDER, PARABOL.BODEN, HALBKUGEL BODEN, FL. BODEN SCHR., PARABOLENDEN, KUGEL, KONISCHER BODEN, KURV.TABELLE, LINEAR.TABELLE Voreinstellung: LINEAR

Bei Auswahl von KURV.TABELLE oder LINEAR.TABELLE geben Sie die Werte fr Fllstand- und Volumensttzpunkte ein, nachdem der Assistent beendet ist. (Siehe Parameter 2.6.7.Tabelle 1-8 in der ausfhrlichen Betriebsanleitung.) Einheit

Maeinheiten des Sensors..

Optionen

M, CM, MM, FT, IN Voreinstellung: M

Hinweis: Fr dieses Beispiel werden alle Werte in Metern (m) angenommen.

7ML19985XU81

SITRANS LUT400 (HART)

Seite DE-21

Oberer Kalibrierungspunkt

Abstand vom Sensor-Bezugspunkt zum Oberen Kalibrierungspunkt: entspricht in der Regel dem Vollpunkt des Prozesses.

Wert

Bereich: 0,000 ... 60,000 Voreinstellung: 0,000

Unterer Kalibrierungspunkt

Abstand vom Sensor-Bezugspunkt zum Unteren Kalibrierungspunkt: entspricht in der Regel dem Nullpunkt des Prozesses.

Wert
Reaktionszeit

Bereich: 0,000 ... 60,000 Voreinstellung: 60,000

Einstellung der Geschwindigkeit, mit der das Gert auf nderungen im Zielbereich reagiert.

Hinweise:

Die Reaktionszeit kann nur ber den Schnellstartassistenten eingestellt werden. Alle nderungen der Parameter Befllgeschwindigkeit pro Minute (2.3.1.) oder Entleergeschwindigkeit pro Minute (2.3.2.), die nach Beenden des Assistenten vorgenommen werden, ersetzen die Einstellung der Reaktionszeit. Reaktionszeit wird immer in m/min angezeigt.
LANGSAM (0,1 M/MIN) MITTEL (1,0 M/MIN) SCHNELL (10 M/MIN) Voreinstellung: LANGSAM (0,1 M/MIN)

Deutsch

Optionen

Die Einstellung sollte die max. Geschwindigkeit beim Befllen oder Entleeren (es gilt der grere Wert) gerade etwas bersteigen. Niedrigere Werte ergeben eine hhere Genauigkeit, whrend hhere Werte schnellere Fllstandschwankungen bercksichtigen knnen. Ma A

Hhe des Behlterbodens in Sensoreinheiten bei einem konischen, pyramiden-, parabol-, kugel-, zylinderfrmigen Boden oder flachen Schrgboden.

Wert
Ma L

Bereich: 0,000 ... 99,999 Voreinstellung: 0,000

Lnge des zylinderfrmigen Teils eines liegenden Behlters mit Parabolenden.

Wert

Bereich: 0,000 ... 99,999 Voreinstellung: 0,000

Volumeneinheiten

Bestimmt die Maeinheiten fr das Volumen.

Optionen

L, USGAL, IMPGAL, M3, BENUTZERDEFINIERT * Voreinstellung: L

* Bei Auswahl der Option BENUTZERDEFINIERT muss der Wert nach Beenden des Assistenten eingestellt werden. Siehe Parameter 2.6.6.Benutzerdefinierte Einheiten in der ausfhrlichen Betriebsanleitung. Seite DE-22 SITRANS LUT400 (HART) 7ML19985XU81

Maximales Volumen

Das maximale Volumen des Behlters. Geben Sie das Behltervolumen ein, das dem Oberen Kalibrierungspunkt entspricht. Beispiel: Bei einem maximalen Behltervolumen von 8000 L geben Sie den Wert 8000 ein.

Wert

Bereich: 0,0 ... 9999999 Voreinstellung: 100,0

Ende des Assistenten QS Volumen

Fr einen erfolgreichen QS mssen alle nderungen angewendet werden.

Optionen Bei Auswahl von ABBRECHEN werden keine nderungen an das Gert
geschrieben.)

ZURCK, ABBRECHEN, FERTIGSTELLEN (Anzeige kehrt auf Men 1.1 Schnellstart zurck, wenn der Schnellstart erfolgreich beendet oder abgebrochen ist.

Um Schnellstart-Werte an das Gert zu bertragen und in das Men der Programmierung zurckzukehren, drcken Sie den Pfeil nach UNTEN cken Sie daraufhin dreimal den LINKS-Pfeil kehren. (Fertigstellen). Dr, um in den Messmodus zurckzu-

1.1.3. QS Durchfluss
Mit diesem Assistenten werden einfache Durchflussapplikationen konfiguriert. [Nur sichtbar auf den Ausfhrungen LUT430 (Pumpen und Durchfluss) und LUT440 (OCM).]

Deutsch

a) Drcken Sie den RECHTS-Pfeil zum Aufruf des PROGRAMMIER-Modus und ffnen Sie die Menebene 1: HAUPTMEN. b) Drcken Sie den RECHTS-Pfeil zweimal, um zum Meneintrag 1.1.1 zu gelangen. , um QS

c) Drcken Sie den Pfeil nach UNTEN zweimal und den RECHTS-Pfeil Durchfluss zu ffnen. d) In jedem Schritt knnen Sie mit dem Pfeil Vorgabewerte akzeptieren nach UNTEN und direkt zum nchsten Eintrag gelangen, oder mit dem RECHTS-Pfeil den Bearbeitungsmodus starten: die aktuelle Auswahl ist hervorgehoben. e) Blttern Sie auf den gewnschten Eintrag und , um die drcken Sie den RECHTS-Pfeil nderung zu speichern, gefolgt vom Pfeil nach UNTEN f) , um fortzufahren. Sie knnen jederzeit den Pfeil nach OBEN LINKS-Pfeil Start des Assistenten QS Durchfluss
SCHNELLSTART DURCHFLUSS
XRS-5
ZURCK ABBRECHEN WEITER BEARBEITEN

ULTRASCHALL-SENSOR

SCHNELLSTART DURCHFLUSS
ULTRASCHALL-SENSOR

KEIN SENSOR XPS-10 XPS-15 XCT-8 XCT-12

XRS-5

drcken, um zurckzufahren, oder den

, um den Assistenten abzubrechen.

Zeigt den Typ des auszufhrenden Assistenten an.

Optionen
7ML19985XU81

ABBRECHEN, START

SITRANS LUT400 (HART)

Seite DE-23

Ultraschall-Sensor

Angabe des angeschlossenen Siemens Sensortyps.

Optionen

KEIN SENSOR, XRS-5, XPS-10, XPS-15, XCT-8, XCT-12, XPS-30, XPS-40, XLT-30, XLT-60, STH Voreinstellung: KEIN SENSOR

Temperaturmessung

Quelle des angezeigten Temperaturwertes fr die Korrektur der Schallgeschwindigkeit.

Optionen

ULTRASCHALLSENSOR, TEMPERATURVORGABE, EXTERNER TS-3, MITTELWERT VON SENSOREN Voreinstellung: ULTRASCHALL-SENSOR

Temperaturvorgabe

Quelle des angezeigten Temperaturwertes fr die Korrektur der Schallgeschwindigkeit.

Wert

Bereich: -100,0 ... +150,0 C Voreinstellung: +20,0 C

Dieser Parameter wird nur angezeigt, wenn TEMPERATURVORGABE in Temperaturmessung gewhlt ist. Messbauwerk

Definiert das in der Applikation anzuwendende Messbauwerk.


EXPONENTIALE MESSBAUWERKE, RECHTECKIGES GERINNE BS-3680, RUNDKRONIGES HORIZ. WEHR BS-3680, TRAPEZF. GERINNE BS-3680, U-PROFIL BS-3680, BREITKRONIGES WEHR BS-3680, DNNWANDIGES RECHTECKIGES WEHR BS-3680, DNNWANDIGES DREIECKSWEHR BS-3680, RECHTECKIGES WEHR EINGEENGT, ROHRPROFIL, PALMERBOWLUSRINNE, H-GERINNE, ANDERE* Voreinstellung: EXPONENTIALE MESSBAUWERKE

Deutsch

Optionen

* Die Option ANDERE wird eingestellt, wenn der Assistent zuvor ber ein HART-Softwaretool (z. B. SIMATIC PDM) betrieben wurde und das Gert auf AUS oder Q/HKENNLINIE (MENGE/HHE) eingestellt wurde. Wenn es sich um die Erstkonfiguration handelt, kann das Messbauwerk nur ber HART-Softwaretools (SIMATIC PDM, AMS, FC375/475) auf AUS (keine Berechnung) oder Linearisierung (Q/H-KENNLINIE (MENGE/HHE)) programmiert werden. Siehe ausfhrliche Betriebsanleitung. Methode der Durchflussberechnung

Bestimmt die Methode fr die Durchflussberechnung.

Optionen
Einheit

ABSOLUT, RATIOMETRISCH Voreinstellung: ABSOLUT

Maeinheiten des Sensors.

Optionen

M, CM, MM, FT, IN Voreinstellung: M

Hinweis: Fr dieses Beispiel werden alle Werte in Metern (m) angenommen.

Seite DE-24

SITRANS LUT400 (HART)

7ML19985XU81

Oberer Kalibrierungspunkt

Abstand vom Sensor-Bezugspunkt zum Oberen Kalibrierungspunkt: entspricht in der Regel dem Vollpunkt des Prozesses.

Wert

Bereich: 0,000 ... 60,000 Voreinstellung: 0,000

Unterer Kalibrierungspunkt

Abstand vom Sensor-Bezugspunkt zum Unteren Kalibrierungspunkt: entspricht in der Regel dem Nullpunkt des Prozesses.

Wert
Reaktionszeit

Bereich: 0,000 ... 60,000 Voreinstellung: 60,000

Einstellung der Geschwindigkeit, mit der das Gert auf nderungen im Zielbereich reagiert.

Hinweise:

Die Reaktionszeit kann nur ber den Schnellstartassistenten eingestellt werden. Alle nderungen der Parameter Befllgeschwindigkeit pro Minute (2.3.1.) oder Entleergeschwindigkeit pro Minute (2.3.2.), die nach Beenden des Assistenten vorgenommen werden, ersetzen die Einstellung der Reaktionszeit. Reaktionszeit wird immer in m/min angezeigt.
LANGSAM (0,1 M/MIN) MITTEL (1,0 M/MIN) SCHNELL (10 M/MIN) Voreinstellung: LANGSAM (0,1 M/MIN)

Optionen

Deutsch

Die Einstellung sollte die max. Geschwindigkeit beim Befllen oder Entleeren (es gilt der grere Wert) gerade etwas bersteigen. Niedrigere Werte ergeben eine hhere Genauigkeit, whrend hhere Werte schnellere Fllstandschwankungen bercksichtigen knnen. Berechnungsfaktoren:

Hinweise:

Die folgenden fnf Parameter erscheinen in dem Assistenten abhngig vom oben gewhlten Messbauwerk. Diese Parameter werden in der Durchflussberechnungsformel eingesetzt (nhere Angaben finden Sie in der Betriebsanleitung). Durchflussexponent (MESSBAUWERK = EXPONENTIAL)

Exponent fr die Berechnungsformel des Durchflusses.

Wert

Bereich: -999,000 ... 9999,000 Voreinstellung: 1,550

7ML19985XU81

SITRANS LUT400 (HART)

Seite DE-25

K-Faktor (MESSBAUWERK = EXPONENTIAL)

Die Konstante zum Einsatz in der Durchflussberechnungsformel, nur fr die absolute Berechnung eines exponentialen Bauwerks.

Wert

Bereich: -999,000 ... 9999,000 Voreinstellung: 1,000

Winkel Dreiecksffnung (MESSBAUWERK = DNNWANDIGES DREIECKSWEHR)

Der Winkel der Dreiecksffnung wird in der Durchflussberechnungsformel eingesetzt.

Wert
Geflle

Bereich: 25,000 ... 95,000 Voreinstellung: 25,000

(MESSBAUWERK = TRAPEZF. GERINNE oder ROHRPROFIL)

Das Durchflussgeflle wird in der Durchflussberechnungsformel eingesetzt.

Wert

Bereich: -999,000 ... 9999,000 Voreinstellung: 0,000

Rauhigkeitskoeffizient (MESSBAUWERK = ROHRPROFIL)

Deutsch

Der Durchfluss-Rauhigkeitskoeffizient wird in der Durchflussberechnungsformel eingesetzt.

Wert

Bereich: -999,000 ... 9999,000 Voreinstellung: 0,000

Seite DE-26

SITRANS LUT400 (HART)

7ML19985XU81

Mae Messbauwerk

Hinweise:

Mit Ausnahme der Optionen Exponentiale Bauwerke und Andere mssen fr jedes Messbauwerk bis zu vier Mae eingegeben werden. Im Assistenten erhalten Sie fr jedes Ma, das fr das gewhlte Bauwerk erforderlich ist, eine Eingabeaufforderung; die jeweilige Mabezeichnung wird angezeigt. Gewhltes Messbauwerk Mabezeichnung im Assistent (Bezug auf Parametermen)
ZULAUFBREITE B (2.15.4.5. Gerinneabmessung 1) EINSCHNRUNGSBREITE B (2.15.4.6. Gerinneabmessung 2) SOHLSCHWELLENHHE P (2.15.4.7. Gerinneabmessung 3) EINSCHNRUNGSLNGE L (2.15.4.8. Gerinneabmessung 4)

Rechteckiges Gerinne BS-3680

Rundkroniges horizontales Wehr BS-3680


SOHLSCHWELLENBREITE B (2.15.4.5. Gerinneabmessung 1) SOHLSCHWELLENHHE P (2.15.4.6. Gerinneabmessung 2) SOHLSCHWELLENLNGE L (2.15.4.7. Gerinneabmessung 3)

Trapezfrmiges Gerinne BS-3680


ZULAUFBREITE B (2.15.4.5. Gerinneabmessung 1) EINSCHNRUNGSBREITE B (2.15.4.6. Gerinneabmessung 2) SOHLSCHWELLENHHE P (2.15.4.7. Gerinneabmessung 3) EINSCHNRUNGSLNGE L (2.15.4.8. Gerinneabmessung 4)

Deutsch

U-Profil BS-3680
ZULAUFDURCHMESSER DA (2.15.4.5. Gerinneabmessung 1) EINSCHNRUNGSDURCHM. D (2.15.4.6. Gerinneabmessung 2) SOHLSCHWELLENHHE P (2.15.4.7. Gerinneabmessung 3) EINSCHNRUNGSLNGE L (2.15.4.8. Gerinneabmessung 4)

Breitkroniges Wehr BS-3680


SOHLSCHWELLENBREITE B (2.15.4.5. Gerinneabmessung 1) SOHLSCHWELLENHHE P (2.15.4.6. Gerinneabmessung 2) SOHLSCHWELLENLNGE L (2.15.4.7. Gerinneabmessung 3)

Dnnwandiges, rechteckiges Wehr BS-3680


ZULAUFBREITE B (2.15.4.5. Gerinneabmessung 1) SOHLSCHWELLENBREITE B (2.15.4.6. Gerinneabmessung 2) SOHLSCHWELLENHHE P (2.15.4.7. Gerinneabmessung 3)

Rechteckiges Wehr eingeengt


SOHLSCHWELLENBREITE B (2.15.4.5. Gerinneabmessung 1) (Fortsetzung auf nchster Seite)

7ML19985XU81

SITRANS LUT400 (HART)

Seite DE-27

Gewhltes Mabezeichnung im Assistent Messbauwerk (Forts.) (Bezug auf Parametermen) Rohrprofil


ROHRINNENDURCHMESSER D (2.15.4.5. Gerinneabmessung 1)

Palmer-Bowlusrinne
MAX. KANALBREITE HMAX (2.15.4.5. Gerinneabmessung 1)

H-Gerinne
MAX. VERZEICHNETE BERFALLHHE HMAX (2.15.4.5. Gerinneabmessung 1)

Max. berfallhhe

Der maximale Fllstandwert, der dem Messbauwerk zugeordnet ist.

Wert

Bereich: 0,000 ... 60,000 Voreinstellung: 60,000

Offset Nullpunkt berfallhhe

Die (positive oder negative) Differenz zwischen Unterer Kalibrierungspunkt und Nullpunkt der berfallhhe (Fllstand bei Null Durchfluss).

Wert

Bereich: -60,000 ... 60,000 Voreinstellung: 0,000

Durchflusseinheiten

Volumeneinheiten fr die Anzeige der Durchflussmenge.

Deutsch

Optionen

L/S, L/MIN, CUFT/S, CUFT/D, GAL/MIN, GAL/D, IMPGAL/MIN, IMPGAL/D, M3/H, M3/TAG, BENUTZERDEFINIERT * Voreinstellung: L/S

* Bei Auswahl der Option BENUTZERDEFINIERT muss der Wert nach Beenden des Assistenten eingestellt werden. Siehe Parameter 2.6.6.Benutzerdefinierte Einheiten in der ausfhrlichen Betriebsanleitung. Maximaler Durchfluss bei 20 mA

Maximale Durchflussmenge.

Wert

Bereich: -999 ... 9999999 Voreinstellung: 100

Dezimalstellen Durchflussmenge

Maximale Anzahl angezeigter Dezimalstellen.

Optionen

KEINE STELLEN, 1 STELLE, 2 STELLEN, 3 STELLEN Voreinstellung: KEINE STELLEN

Min. Mengenunterdrckung Durchfluss

Bei berfallhhen am oder unterhalb des Minimalwertes erfolgt keine Summierung.

Wert

Bereich: 0,000 ... 60,000 Voreinstellung: 0,000

Seite DE-28

SITRANS LUT400 (HART)

7ML19985XU81

Ende des QS-Assistenten Durchfluss

Fr einen erfolgreichen QS mssen alle nderungen angewendet werden.

Optionen

ZURCK, ABBRECHEN, FERTIGSTELLEN (Anzeige kehrt auf Men 1.1.1 Schnellstart zurck, wenn der Schnellstart erfolgreich beendet oder abgebrochen ist. Bei Auswahl von ABBRECHEN werden keine nderungen an das Gert geschrieben.)

Um Schnellstart-Werte an das Gert zu bertragen und in das Men der Programmierung zurckzukehren, drcken Sie den Pfeil nach UNTEN cken Sie daraufhin dreimal den LINKS-Pfeil zurckzukehren. (Fertigstellen). Dr, um in den Messmodus

Hinweis: Fr eine optimale Genauigkeit ist es sehr empfehlenswert, nach Beenden des Assistenten eine 2.15.2.Autom. Nullpunktkorrektur berfallhhe durchzufhren. Nhere Angaben in der Betriebsanleitung.

1.2. Pumpensteuerung
Mit diesem Assistenten knnen Pumpen konfiguriert werden, wenn sie in Ihrer Applikation verwendet werden. Fhren Sie zuerst den zutreffenden Schnellstartassistenten aus. a) Drcken Sie den RECHTS-Pfeil zum Aufruf des PROGRAMMIER-Modus und ffnen Sie die Menebene 1: HAUPTMEN. b) Drcken Sie den RECHTS-Pfeil c) Drcken Sie den Pfeil nach UNTEN Pumpensteuerung zu ffnen. d) In jedem Schritt knnen Sie mit dem Pfeil nach UNTEN akzeptieren und direkt zum nchsten Eintrag gelangen, oder mit dem RECHTS-Pfeil Auswahl ist hervorgehoben. Vorgabewerte , um zum Meneintrag 1.1 zu gelangen. und den RECHTS-Pfeil , um die

Deutsch

den Bearbeitungsmodus starten: die aktuelle ,

e) Blttern Sie auf den gewnschten Eintrag und drcken Sie den RECHTS-Pfeil um die nderung zu speichern, gefolgt vom Pfeil nach UNTEN f) Sie knnen jederzeit den Pfeil nach OBEN LINKS-Pfeil Start des Assistenten - Pumpensteuerung

, um fortzufahren.

drcken, um zurckzufahren, oder den

, um den Assistenten abzubrechen.

Zeigt den Typ des auszufhrenden Assistenten an.

Optionen
Anzahl Pumpen

ABBRECHEN, START

Auswahl der Anzahl Pumpen, die im Zusammenhang mit der Pumpensteuerung verwendet werden.

Optionen

OHNE, 2 Voreinstellung: OHNE

Bei Auswahl von OHNE ist die Pumpensteuerung deaktiviert. 7ML19985XU81 SITRANS LUT400 (HART) Seite DE-29

Relais Pumpe 1

Auswahl des Relais, das Pumpe 1 zugeordnet ist.

Optionen
Relais Pumpe 2

RELAIS 2, RELAIS 3 Voreinstellung: RELAIS 2

Nur zur Ansicht. Automatische Einstellung des Relais, das Pumpe 2 zugeordnet ist, basierend auf dem fr Pumpe 1 gewhlten Relais im vorigen Schritt.

Optionen (nur zur Ansicht)

Wenn Relais Pumpe 1 = RELAIS 2, dann ist Relais Pumpe 2 = RELAIS 3 Wenn Relais Pumpe 1 = RELAIS 3, dann ist Relais Pumpe 2 = RELAIS 2

Modus Pumpensteuerung

Steueralgorithmus zum Auslsen des Relais.

Auswertegert SITRANS LUT420 Fllstand:


STAFFEL MIT VERTAUSCHUNG, ERSATZBETRIEB MIT VERTAUSCHUNG

Auswertegert LUT430 Pumpen und Durchfluss: Je nach Ausfhrung untersttzte SITRANS LUT440 OCM (hochgenaue DurchflussmesOptionen sung): Deutsch
STAFFEL MIT VERTAUSCHUNG, ERSATZBETRIEB MIT VERTAUSCHUNG, NUTZUNGSVERHLTNIS STAFFEL, NUTZUNGSVERHLTNIS ERSATZBETRIEB, STAFFEL OHNE VERTAUSCHUNG, ERSATZBETRIEB OHNE VERTAUSCHUNG

STAFFEL MIT VERTAUSCHUNG, ERSATZBETRIEB MIT VERTAUSCHUNG, NUTZUNGSVERHLTNIS STAFFEL, NUTZUNGSVERHLTNIS ERSATZBETRIEB, STAFFEL OHNE VERTAUSCHUNG, ERSATZBETRIEB OHNE VERTAUSCHUNG Voreinstellung (alle Ausfhrungen): STAFFEL MIT VERTAUSCHUNG

Nutzungsverhltnis Pumpe 1

Fr die Auswahl einer Pumpe ist die Laufzeit ausschlaggebend und nicht, welche Pumpe zuletzt verwendet wurde.

Wert

Bereich: 0 ... 255 Voreinstellung: 1

Dieser Parameter wird nur angezeigt, wenn ein Algorithmus Nutzungsverhltnis in Modus Pumpensteuerung gewhlt wurde. Nutzungsverhltnis Pumpe 2

Fr die Auswahl einer Pumpe ist die Laufzeit ausschlaggebend und nicht, welche Pumpe zuletzt verwendet wurde.

Wert

Bereich: 0 ... 255 Voreinstellung: 1

Dieser Parameter wird nur angezeigt, wenn ein Algorithmus Nutzungsverhltnis in Modus Pumpensteuerung gewhlt wurde.

Seite DE-30

SITRANS LUT400 (HART)

7ML19985XU81

Betriebsdauer Relais 2

Einstellung der Zeit, die Pumpenrelais 2 in Betrieb war; in Stunden.

Wert

Bereich: 0 ... 999999 Voreinstellung: 0

Verwenden Sie den Vorgabewert fr neue Pumpen, oder fr bestehende Pumpen mit kumulierter Betriebsdauer. (Dieser Parameter wird nur angezeigt, wenn ein Algorithmus Nutzungsverhltnis in Modus Pumpensteuerung gewhlt wurde.) Betriebsdauer Relais 3

Einstellung der Zeit, die Pumpenrelais 3 in Betrieb war; in Stunden.

Wert

Bereich: 0 ... 999999 Voreinstellung: 0

Verwenden Sie den Vorgabewert fr neue Pumpen, oder fr bestehende Pumpen mit kumulierter Betriebsdauer. (Dieser Parameter wird nur angezeigt, wenn ein Algorithmus Nutzungsverhltnis in Modus Pumpensteuerung gewhlt wurde.) EIN-Schaltpunkt Pumpe 1

Der Fllstand, an dem Pumpe 1 EIN-schaltet, definiert in 2.1.1.Einheit.

Wert

Bereich: 0,000 ... 99999,000 Voreinstellung: 0,000

EIN-Schaltpunkt Pumpe 2

Der Fllstand, an dem Pumpe 2 EIN-schaltet, definiert in 2.1.1.Einheit.

Deutsch

Wert

Bereich: 0,000 ... 99999,000 Voreinstellung: 0,000

AUS-Schaltpunkt Pumpe 1

Der Fllstand, an dem Pumpe 1 AUS-schaltet, definiert in 2.1.1.Einheit.

Wert

Bereich: 0,000 ... 99999,000 Voreinstellung: 0,000

AUS-Schaltpunkt Pumpe 2

Der Fllstand, an dem Pumpe 2 AUS-schaltet, definiert in 2.1.1.Einheit.

Wert

Bereich: 0,000 ... 99999,000 Voreinstellung: 0,000

Ende des Assistenten - Pumpensteuerung

Fr einen erfolgreich beendeten Assistenten mssen alle nderungen angewendet werden.

Optionen

ZURCK, ABBRECHEN, FERTIGSTELLEN (Anzeige kehrt auf Men Pumpensteuerung zurck, wenn der Assistent erfolgreich beendet oder abgebrochen ist. Bei Auswahl von ABBRECHEN werden keine nderungen an das Gert geschrieben.)

Um Werte an das Gert zu bertragen und in das Men der Programmierung zurckzukehren, drcken Sie den Pfeil nach UNTEN (Fertigstellen). Drcken Sie daraufhin zweimal den LINKS-Pfeil , um in den Messmodus zurckzukehren. 7ML19985XU81 SITRANS LUT400 (HART) Seite DE-31

SITRANS LUT400 Kommunikation: HART


SITRANS LUT400 kann ber das HART-Netzwerk konfiguriert werden, und zwar mit dem HART Communicator 375 von Emerson (nhere Angaben finden Sie in der Betriebsanleitung) oder mit einem Softwarepaket. Empfohlen wird das Softwarepaket SIMATIC Process Device Manager (PDM) von Siemens. Das Applikationsbeispiel Working with HART liefert nhere Angaben und ist zum Download auf der Produktseite unserer Website verfgbar. Siehe: www.siemens.de/sitransLUT400 unter Support und klicken Sie auf Applikationsbeispiele.

Fernzugriff
SITRANS LUT400 untersttzt mehrere Softwaretools fr einen Fernzugriff ber Kommunikation: PC mit SIMATIC PDM PC mit Emerson AMS Device Manager PC mit Webbrowser PC mit Field Device Tool (FDT) Field Communicator 375/475 (FC375/FC475). Untenstehend finden Sie allgemeine Informationen zur Verwendung dieser Tools mit Ihrem SITRANS LUT400. Nhere Angaben zu jedem Tool enthlt das Handbuch Kommunikation fr SITRANS LUT400 (HART).

SIMATIC PDM
SIMATIC PDM ist ein Softwarepaket fr die Inbetriebnahme und Wartung des SITRANS LUT400 und anderer Prozessgerte. Weitere Informationen finden Sie unter www.siemens.com/simatic-pdm. Prfen Sie die Support-Seite unserer Website, um sicherzustellen, dass Sie die neueste Version von SIMATIC PDM, das aktuellste Servicepaket (SP) und den aktuellsten Hotfix (HF) haben. Siehe: http://support.automation.siemens.com/WW/ llisapi.dll?func=cslib.csinfo&lang=en&objiD=10806857&subtype=133100

Deutsch

Aktualisieren der Electronic Device Description (EDD)


Sie finden die EDD im Gertekatalog, unter Sensors/Level/Echo/Siemens AG/SITRANS LUT400. (Die EDD ist fr eine Vorwrtskompatibilitt ausgelegt.) Als Richtlinie zur Festlegung der richtigen EDD achten Sie darauf, dass die Major- und Minor-Nummern von EDD-Version und FirmwareVersion im Gert bereinstimmen (z. B. Major- und Minor-Nummern fettgedruckt: 1.00.00-02). Die Revisionsnummer der EDD muss mit der Firmware-Version des Gertes bereinstimmen. Um dies in PDM zu berprfen, gehen Sie zu SITRANS LUT400 HART > Identifikation > Gert.

bereinstimmende Firmwareund EDDVersion

Um eine neue Version von SIMATIC PDM zu installieren, mssen Sie das neueste Service Pack (SP) und den neuesten Hotfix (HF) haben.

Installieren einer neuen EDD

Gehen Sie zu www.siemens.com/sitransLUT400 > Support > Software Downloads, um die aktuellste EDD von der Produktseite unserer Website herunterzuladen. Speichern Sie die Dateien auf Ihrem Computer und extrahieren Sie die komprimierte Datei an eine leicht erreichbare Stelle. Starten Sie SIMATIC PDM Gertekatalog verwalten, blttern Sie bis zum Ordner, der die entzippte EDD-Datei enthlt, und whlen Sie diesen. SITRANS LUT400 (HART) 7ML19985XU81

Seite DE-32

Konfigurieren eines neuen Gertes


1) 2) berprfen Sie, ob Sie die neueste EDD besitzen und aktualisieren Sie sie bei Bedarf (siehe Installieren einer neuen EDD weiter oben). Konfigurieren Sie SITRANS LUT400 mit dem Schnellstartassistenten fr Ihre Applikation.

Schnellstartassistenten ber SIMATIC PDM


Hinweise:
Die Einstellungen des Schnellstartassistenten sind zusammenhngend und nderungen werden erst wirksam, wenn Sie am Ende des letzten Schrittes auf bernehmen und bertragen klicken. Klicken Sie auf ZURCK, um zurckzukehren und die Einstellung zu berprfen oder auf ABBRECHEN, um den Schnellstart zu verlassen. ffnen Sie das Men Gert in SIMATIC PDM und whlen Sie den Schnellstart, der auf Ihre Applikation zutrifft (Fllstand, Volumen, Volumen-Linearisierung, Durchfluss).

Schnellstart

Betrieb ber AMS Device Manager


AMS Device Manager ist ein Softwarepaket fr die Inbetriebnahme und Wartung des SITRANS LUT400 und anderer Prozessgerte. Weitere Informationen finden Sie unter: http://www.emersonprocess.com/AMS/.

Deutsch

Electronic Device Description (EDD)


SITRANS LUT400 erfordert die EDD fr den AMS Device Manager Version 10.5.

Konfigurieren eines neuen Gerts ber AMS Device Manager


1) Sie finden die EDD im Gertekatalog, unter Sensors/Level/Echo/Siemens AG/SITRANS LUT400. Gehen Sie zur Produktseite unserer Website unter: www.siemens.com/sitransLUT400, unter Support->Software Downloads, und prfen Sie, dass Sie die neueste Ausfhrung der EDD fr AMS Device Manager haben. Speichern Sie die Dateien auf Ihrem Computer und extrahieren Sie die komprimierte Datei an eine leicht erreichbare Stelle. Starten Sie AMS Device Manger Add Device Type (Gertetyp hinzufgen), blttern Sie bis zur entzippten EDD-Datei und whlen Sie diese. (Ein Applikationsbeispiel zur Einstellung von HART-Gerten mit AMS Device Manager kann unter Support von der Produktseite unserer Website heruntergeladen werden.) In der Device Connection View (Ansicht Gerteanschluss) klicken Sie mit der rechten Maustaste auf das Gertesymbol und whlen Sie Scan Device (Gerte-Scan), um Parameter vom Gert zu laden. Mit einem Doppelklick auf das Gertesymbol wird der Anlauf-Bildschirm geffnet. Der Anlauf-Bildschirm zeigt die Details der Gerteidentifikation und ein Navigationsfenster links im Bildschirm. Konfigurieren Sie SITRANS LUT400 mit dem geeigneten Schnellstartassistenten. SITRANS LUT400 (HART) Seite DE-33

2) 3)

4)

5)

6)

7ML19985XU81

Betrieb ber FC375/475 (HART)


Der FC375/FC475 HART Communicator ist ein bedienerfreundliches Handheld-Kommunikationsgert. Es bietet eine universelle Untersttzung fr HART-Gerte, wie SITRANS LUT400. Die FC375/475-Menstruktur ist der Menstruktur fr AMS Device Manager sehr hnlich.

Electronic Device Description (EDD)


SITRANS LUT400 erfordert die EDD fr FC375/FC475.

Konfigurieren eines neuen Gerts ber FC375/FC475


1) Gehen Sie zur Produktseite unserer Website: www.siemens.com/sitransLUT400, unter Support->Software Downloads, und prfen Sie, dass Sie die neueste Ausfhrung der EDD fr FC375/FC475 haben. Speichern Sie die Dateien auf Ihrem Computer und extrahieren Sie die komprimierte Datei an eine leicht erreichbare Stelle. Zur Installation der EDD lassen Sie die Software Easy Upgrade Utility des Field Communicator von Emerson laufen. Konfigurieren Sie SITRANS LUT400 mit dem Schnellstartassistenten fr Ihre Applikation.

2) 3) 4)

Bedienung ber FDT (Field Device Tool)


FDT ist ein in mehreren Softwarepaketen verwendeter Standard fr die Inbetriebnahme und Wartung von Feldgerten, wie z. B. SITRANS LUT400. PACTware und Fieldcare sind zwei handelsbliche FDTs. FDT und PDM sind sich sehr hnlich. Um ein Feldgert ber FDT zu projektieren, ist der DTM (Device Type Manager) fr dieses Gert erforderlich. In Siemens Messgerten wird SITRANS DTM, sowie eine fr SITRANS DTM geschriebene EDD eingesetzt. 1) Installieren Sie zunchst SITRANS DTM auf Ihrem System. Der Download ist unter folgender Adresse mglich: http://www.siemens.com/sitransdtm. Klicken Sie auf Support, und dann auf Software Downloads. 2) Installieren Sie die SITRANS LUT400 (HART) EDD fr SITRANS DTM. Diese steht auf der Produktseite unserer Website zum Download zur Verfgung. Siehe www.siemens.de/sitransLUT400 unter Support und klicken Sie auf Software Downloads.

Deutsch

Konfiguration eines neuen Gerts ber FDT


Applikationsbeispiele stehen auf der Produktseite unserer Website unter Support zum Download zur Verfgung.

Instandhaltung und Wartung


Unter normalen Betriebsbedingungen erfordert SITRANS LUT400 keine Wartung oder Reinigung.

Gertereparatur und Haftungsausschluss


Nhere Angaben finden Sie auf der hinteren Umschlaginnenseite.

Seite DE-34

SITRANS LUT400 (HART)

7ML19985XU81

Anschlussmethoden bei Einbau in explosionsgefhrdeten Bereichen


Fr Installationen in Ex-Bereichen stehen die folgenden Anschlussoptionen zur Auswahl: Anschlussmethode Staubexplosionssicher (International/Europa) Anschlussmethode Nicht zndgefhrlich/Non-incendive (Kanada) Prfen Sie in allen Fllen die Zulassungen auf dem Typenschild Ihres Gertes.

1. Anschlussmethode Staubexplosionssicher (International/Europa)


SITRANS LUT420 7MLxxxx-xxxxx-xxxx Serial No.: GYZ / B1034567 0518 KCC-REM-S49 Power Rating: 100 230V 15% SITRANSLUT400 50/60 Hz, 36 VA (10 W) Contact Rating: 1A/5A @ 250V , Non-Inductive Operating Temperature: 20C to 50C Enclosure: IP65 / TYPE 4X / NEMA 4X II 3D Ex tc IIIC T70C Dc Ex tc IIIC T70C Dc SIRA 12ATEX9143X IECEx SIR 12.0057X Warning: Possible static hazard. Do not rub or clean on site with a dry cloth. Warning: Do not open when energized or when an explosive dust atmosphere is present. Siemens Milltronics Process Instruments, Peterborough Assembled in Canada with domestic and imported parts SITRANS LUT430 7MLxxxx-xxxxx-xxxx Serial No.: GYZ / B1034567 0518 KCC-REM-S49 Power Rating: 100 230V 15% SITRANSLUT400 50/60 Hz, 36 VA (10 W) Contact Rating: 1A/5A @ 250V , Non-Inductive Operating Temperature: 20C to 50C Enclosure: IP65 / TYPE 4X / NEMA 4X II 3D Ex tc IIIC T70C Dc Ex tc IIIC T70C Dc SIRA 12ATEX9143X IECEx SIR 12.0057X Warning: Possible static hazard. Do not rub or clean on site with a dry cloth. Warning: Do not open when energized or when an explosive dust atmosphere is present. Siemens Milltronics Process Instruments, Peterborough Assembled in Canada with domestic and imported parts SITRANS LUT440 7MLxxxx-xxxxx-xxxx Serial No.: GYZ / B1034567 0518 KCC-REM-S49 Power Rating: 100 230V 15% SITRANSLUT400 50/60 Hz, 36 VA (10 W) Contact Rating: 1A/5A @ 250V , Non-Inductive Operating Temperature: 20C to 50C Enclosure: IP65 / TYPE 4X / NEMA 4X II 3D Ex tc IIIC T70C Dc Ex tc IIIC T70C Dc SIRA 12ATEX9143X IECEx SIR 12.0057X Warning: Possible static hazard. Do not rub or clean on site with a dry cloth. Warning: Do not open when energized or when an explosive dust atmosphere is present. Siemens Milltronics Process Instruments, Peterborough Assembled in Canada with domestic and imported parts

SITRANS LUT420 7MLxxxx-xxxxx-xxxx Serial No.: GYZ / B1034567 0518 KCC-REM-S49 Power Rating: 10 32V , 10W SITRANSLUT400 Contact Rating: 1A/5A @ 250V , Non-Inductive Operating Temperature: 20C to 50C Enclosure: IP65 / TYPE 4X / NEMA 4X

SITRANS LUT430 7MLxxxx-xxxxx-xxxx Serial No.: GYZ / B1034567 0518 KCC-REM-S49 Power Rating: 10 32V , 10W SITRANSLUT400 Contact Rating: 1A/5A @ 250V , Non-Inductive Operating Temperature: 20C to 50C Enclosure: IP65 / TYPE 4X / NEMA 4X

SITRANS LUT440 7MLxxxx-xxxxx-xxxx Serial No.: GYZ / B1034567 0518 KCC-REM-S49 Power Rating: 10 32V , 10W SITRANSLUT400 Contact Rating: 1A/5A @ 250V , Non-Inductive Operating Temperature: 20C to 50C Enclosure: IP65 / TYPE 4X / NEMA 4X

II 3D Ex tc IIIC T70C Dc Ex tc IIIC T70C Dc SIRA 12ATEX9143X IECEx SIR 12.0057X Warning: Possible static hazard. Do not rub or clean on site with a dry cloth. Warning: Do not open when energized or when an explosive dust atmosphere is present.

II 3D Ex tc IIIC T70C Dc Ex tc IIIC T70C Dc SIRA 12ATEX9143X IECEx SIR 12.0057X Warning: Possible static hazard. Do not rub or clean on site with a dry cloth. Warning: Do not open when energized or when an explosive dust atmosphere is present.

II 3D Ex tc IIIC T70C Dc Ex tc IIIC T70C Dc SIRA 12ATEX9143X IECEx SIR 12.0057X Warning: Possible static hazard. Do not rub or clean on site with a dry cloth. Warning: Do not open when energized or when an explosive dust atmosphere is present.

Deutsch

Siemens Milltronics Process Instruments, Peterborough Assembled in Canada with domestic and imported parts

Siemens Milltronics Process Instruments, Peterborough Assembled in Canada with domestic and imported parts

Siemens Milltronics Process Instruments, Peterborough Assembled in Canada with domestic and imported parts

Das auf dem Typenschild des Gertes aufgefhrte ATEX-Zertifikat steht auf der Produkt- seite unserer Website zum Download zur Verfgung www.siemens.com/sitransLUT400. Das auf dem Typenschild des Gertes aufgefhrte IECEx-Zertifikat finden Sie auf der IECExWebsite. Siehe: http://iecex.iec.ch, klicken Sie auf Ex Equipment Certificates of Conformity und geben die Zertifikatsnummer IECEx SIR 12.0057X ein. Fr Anschlussanforderungen beachten Sie lokale Richtlinien. Siehe auch das oben aufgefhrte ATEX-Zertifikat.

7ML19985XU81

SITRANS LUT400 (HART)

Seite DE-35

2. Anschlussmethode Nicht zndgefhrlich/Non-incendive (Kanada)


SITRANS LUT420 7MLxxxx-xxxxx-xxxx Serial No.: GYZ / B1034567 Power Rating: 100 230V 15% 50/60 Hz, 36 VA (10 W) Contact Rating: 1A/5A @ 250V , Non-Inductive Operating Temperature: 20C to 50C Enclosure: IP65 / TYPE 4X / NEMA 4X SITRANS LUT430 7MLxxxx-xxxxx-xxxx Serial No.: GYZ / B1034567 Power Rating: 100 230V 15% 50/60 Hz, 36 VA (10 W) Contact Rating: 1A/5A @ 250V , Non-Inductive Operating Temperature: 20C to 50C Enclosure: IP65 / TYPE 4X / NEMA 4X SITRANS LUT440 7MLxxxx-xxxxx-xxxx Serial No.: GYZ / B1034567 Power Rating: 100 230V 15% 50/60 Hz, 36 VA (10 W) Contact Rating: 1A/5A @ 250V , Non-Inductive Operating Temperature: 20C to 50C Enclosure: IP65 / TYPE 4X / NEMA 4X

159134

Class I, Div.2, Gr. A, B, C & D T3C Class II, Div. 2, Gr. F & G Class III Per Drawing A5E03936871

159134

Class I, Div.2, Gr. A, B, C & D T3C Class II, Div. 2, Gr. F & G Class III Per Drawing A5E03936871

159134

Class I, Div.2, Gr. A, B, C & D T3C Class II, Div. 2, Gr. F & G Class III Per Drawing A5E03936871

Siemens Milltronics Process Instruments, Peterborough Assembled in Canada with domestic and imported parts

Siemens Milltronics Process Instruments, Peterborough Assembled in Canada with domestic and imported parts

Siemens Milltronics Process Instruments, Peterborough Assembled in Canada with domestic and imported parts

SITRANS LUT420 7MLxxxx-xxxxx-xxxx Serial No.: GYZ / B1034567 Power Rating: 10 32V , 10W Contact Rating: 1A/5A @ 250V , Non-Inductive Operating Temperature: 20C to 50C Enclosure: IP65 / TYPE 4X / NEMA 4X

SITRANS LUT430 7MLxxxx-xxxxx-xxxx Serial No.: GYZ / B1034567 Power Rating: 10 32V , 10W Contact Rating: 1A/5A @ 250V , Non-Inductive Operating Temperature: 20C to 50C Enclosure: IP65 / TYPE 4X / NEMA 4X

SITRANS LUT440 7MLxxxx-xxxxx-xxxx Serial No.: GYZ / B1034567 Power Rating: 10 32V , 10W Contact Rating: 1A/5A @ 250V , Non-Inductive Operating Temperature: 20C to 50C Enclosure: IP65 / TYPE 4X / NEMA 4X

159134

Class I, Div.2, Gr. A, B, C & D T3C Class II, Div. 2, Gr. F & G Class III Per DWG. A5E03936871

159134

Class I, Div.2, Gr. A, B, C & D T3C Class II, Div. 2, Gr. F & G Class III Per DWG. A5E03936871

159134

Class I, Div.2, Gr. A, B, C & D T3C Class II, Div. 2, Gr. F & G Class III Per DWG. A5E03936871

Siemens Milltronics Process Instruments, Peterborough Assembled in Canada with domestic and imported parts

Siemens Milltronics Process Instruments, Peterborough Assembled in Canada with domestic and imported parts

Siemens Milltronics Process Instruments, Peterborough Assembled in Canada with domestic and imported parts

Die Anschlusszeichnung CSA Class I, Div 2 Nr. A5E03936871 steht auf der Produktseite unserer Website zum Download zur Verfgung: www.siemens.com/sitransLUT400.

Deutsch

Vorschriften bezglich Installationen in explosionsgefhrdeten Bereichen (Europische ATEX Richtlinie 94/9/EG, Anhang II, 1/0/6)
Folgende Vorschriften finden Anwendung auf die Gerte, die Gegenstand des Zertifikats Nr. SIRA 12ATEX9143X sind: 1) Angaben zu Verwendung und Zusammenbau finden Sie im Hauptteil der Vorschriften. 2) Das Gert ist fr den Einsatz als Betriebsmittel der Kategorie 3D zertifiziert. 3) Das Betriebsmittel kann in Ex-Zone 22 eingesetzt werden. 4) Das Betriebsmittel erreicht eine maximale Oberflchentemperatur von T70 C (bei einer Umgebungstemperatur von +50 C). Beziehen Sie sich auf die geltenden Verfahrensregeln fr die Auswahl dieses Betriebsmittels hinsichtlich der spezifischen Temperaturen zur Staubexplosion. 5) Die Zulassung dieser Betriebsmittel gilt fr den Einsatz bei Umgebungstemperaturen von -20 bis +50 C. 6) Alle Kabel- oder Rohreinfhrungen mssen die Anforderungen der Europischen Richtlinie 94/9/EG fr Gruppe II, Kategorie 3D erfllen und die allgemeine Schutzart IP des Gehuses gewhrleisten. 7) Die Auswahl der Kabel sollte in bereinstimmung mit den geltenden Verfahrensregeln erfolgen und die Isolierung muss der maximalen Oberflchentemperatur des Gehuses (T70 C) widerstehen knnen. 8) Das Gert wird nicht als Sicherheitseinrichtung (im Sinne der Richtlinie 94/9/EG Anhang II, Klausel 1,5) eingestuft. 9) Installation und Prfung dieses Gerts drfen nur durch entsprechend geschultes Personal in bereinstimmung mit den geltenden Verfahrensregeln (EN 60079-14 und EN 6007917 in Europa) durchgefhrt werden. 10) Die Reparatur dieses Gerts darf nur durch entsprechend geschultes Personal in bereinstimmung mit den geltenden Verfahrensregeln (z. B. EN 60079-19 in Europa) durchgefhrt werden. Seite DE-36 SITRANS LUT400 (HART) 7ML19985XU81

11) Ins Gert einzubauende oder als Ersatzteil zu verwendende Werkstcke mssen durch entsprechend geschultes Personal in bereinstimmung mit der Dokumentation des Herstellers montiert werden. 12) Ein manuelles Ausschalten ist mit dem Leitungsschutzschalter mglich, der kundenseitig installiert ist.

SONDERBEDINGUNGEN FR SICHERE VERWENDUNG


Die Endung X der Zertifizierungsnummer bezieht sich auf folgende Sonderbedingungen fr eine sichere Verwendung: Alle Kabel- oder Rohreinfhrungen sollen die Anforderungen der Europischen Richtlinie 94/9/EG fr Gruppe II, Kategorie 3D erfllen und die allgemeine Schutzart IP des Gehuses gewhrleisten. Die Auswahl der Kabel soll in bereinstimmung mit den geltenden Verfahrensregeln erfolgen und die Isolierung soll der maximalen Oberflchentemperatur des Gehuses (T70 C) widerstehen knnen. Das Gehuse ist nichtleitend und kann, unter bestimmten, extremen Bedingungen, eine zndfhige elektrische Ladung erzeugen. Der Benutzer sollte sicherstellen, dass die Einbaustelle des Gerts keinen externen Bedingungen (wie z. B. Hochdruckdampf) ausgesetzt ist, welche die Entwicklung elektrostatischer Ladungen auf nichtleitenden Oberflchen verursachen knnten. Das Betriebsmittel sollte auerdem nur mit einem feuchten Lappen gereinigt werden. Das Betriebsmittel sollte entweder in einem Bereich installiert werden, in dem die Gefahr mechanischer Beschdigung gering ist, oder, falls dies nicht mglich ist, durch eine zustzliche Schutzeinrichtung gesichert werden, die einen der Applikation angemessenen mechanischen Schutz bietet. Alle Auswertegerte der Baureihe LUT400 (LUT420, LUT430, LUT440) mit abgesetzt montiertem Deckel (Schalttafeleinbau) sollen in einem Gehuse geeigneter Schutzart installiert werden, welches die Anforderungen der Europischen Richtlinie 94/9/EG fr Gruppe II, Kategorie 3D und mindestens die Schutzart IP65 erfllt.

Deutsch

7ML19985XU81

SITRANS LUT400 (HART)

Seite DE-37

Notizen

Deutsch
Seite DE-38

SITRANS LUT400 (HART)

7ML19985XU81

SITRANS LUT400 (HART)


SITRANS LUT400 (HART 1)). , . SITRANS LUT400 (HART) SITRANS LUT400 (HART), Web : www.siemens.com/sitransLUT400. Siemens. : Siemens AG Siemens Milltronics Process Instruments 1954 Technology Drive, P.O. Box 4225 Peterborough, Ontario, Canada, K9J 7B1 Email: techpubs.smpi@siemens.com
Siemens AG 2012. Siemens Milltronics Process Instruments. Siemens Milltronics Process Instruments , . , . , . . . .

ILLTRONICS Siemens Milltronics Process Instruments.


24 . , Siemens, : www.siemens.com/automation/partner: Contact () ( : Automation Technology ( )). . : +Automation Technology > +Sensor Systems ( ) >+Process Instrumentation ( ) > +Level Measurement ( ) > +Continuous (). Ultrasonic (). : +Automation Technology > +Sensor Systems >+Process Instrumentation > +Flow Measurement ( ). Ultrasonic. /. Technical Support ( ) Service (). on-line : www.siemens.com/automation/ support-request (SITRANS LUT400) , Search () . Next (). . , Next Siemens. IA/DT Siemens: +49 (0)911 895 7222

1)

HART HART Communication Foundation.

7ML19985XU81

SITRANS LUT400 (HART)

EL-1


. . , , / . , , / . : .

FCC
...: (FCC)
: Siemens Milltronics . : A, 15 FCC. , . , , , . , .

EL-2

SITRANS LUT400 (HART)

7ML19985XU81

SITRANS LUT400
: SITRANS LUT400 , . : . . SITRANS LUT400 Siemens , , , , . EchoMax Siemens, 0,3 60 ( ). SITRANS LUT400 LUI ( ) . plug and play. SITRANS LUT400 HART [SIMATIC PDM, AMS Device Manager, Field Communicator 375/475 Field Device Tools (FDTs)]. , Sonic Intelligence, . , SITRANS LUT400 (HART).

AC

1)

DC
10-32 V DC, 10 W 1 : 5 x 20 mm, , 1,6A, 125V

100-230 V AC 15%, 50 / 60 Hz, 36 VA (10 W)1) : 5 x 20 mm, , 0,25 A, 250 V


.

2000 m . 4


-20 +50 C (-4 +122 F)


II


( IP65 / 4X / NEMA 4X)

: . IP NEMA. () CSAUS/C, CE, FM, UL listed, C-TICK

CSA I, . 2, A, B, C, D, II, . 2, F, G, III ATEX II 3D (/) Ex tc IIIC T70 C IP65 IECEx SIR 12.0057X SIRA 12ATEX9143X

7ML19985XU81

SITRANS LUT400 (HART)

EL-3


: . . . 250 V. . DC , IEC 61010-1. 250 V. 250 V. . .


, , DIN . . : , 9 SITRANS LUT400.


SITRANS LUT400 . LUI .

(2 ) (2 )

: .

EL-4

SITRANS LUT400 (HART)

7ML19985XU81

(2 ) (2 )

: .


1) 2) 3) ( ). . . LUT400 .

: M8 5/16" 17 mm 5/8" : . , .

7ML19985XU81

SITRANS LUT400 (HART)

EL-5

SITRANS LUT400


: 5 m ( 2,5 m ). . 1) LUI . , . . (). . : : 1,1 N m 10 in-lbs

2)

:
: , 5 m . , . : : M5, , M5 . IP65 .

EL-6

SITRANS LUT400 (HART)

7ML19985XU81

: .


1) 2) 3) , ( ) . . , . LUT400 . ( 3 5). : . , . : : 3", : 3/4" : : : 3" M8 3/8" : 3/4" M6 1/4" : M6 1/4" M8 3/8" : : 16 mm 13/16".

7ML19985XU81

SITRANS LUT400 (HART)

EL-7

DIN

DIN

(2)

DIN


1) 2) 3) DIN . DIN DIN. DIN DIN, DIN. : DIN. DIN: TH 35-7,5 TH 35-15 IEC 60715. DIN SITRANS LUT400.

1) DIN . DIN. , DIN.

2)

EL-8

SITRANS LUT400 (HART)

7ML19985XU81


. 1 5 . 1) 2) 3) 4) 5) . (. ). . . . .
SITRANS LUT400

(3 )

(3 )

: , 21,4 mm 21,6 mm. M20 ( 20 mm) 1/2" NPT ( 21,3). . ( O). , IP65 .


( 1 5) 6) 1 5, . ( 1/2" NPT.) 7) .


( 1 5) 6) 1 5, . 7) . . 8) . 9) .

7ML19985XU81

SITRANS LUT400 (HART)

EL-9



, . . LUT400 ( ). , . , .

SITRANS LUT400

: . IP NEMA. : . : 1) 2) 1/4 . . . . , . .

3)

4) 5)

6)

EL-10

SITRANS LUT400 (HART)

7ML19985XU81


LUT400 . . : . , . : . 0,56 - 0,79 N m 5 - 7 in-lbs .

TS-3 TRANSDUCER DISCRETE INPUTS INPUT POWER

8 1

9 2

10

11 0.25 A, 250 V SLOW BLOW

12 L1

13 L2/N 100 - 230 V

14

DC SYNC

SHLD

WHT

BLK

BLK

WHT


mA OUT/HART BATTERY BR2032

RELAYS

DISPLAY 15 16 RLY 1 17 18 19 20 21 22 23 24

RLY 2

RLY 3 ACT

USB
PAS SERVICE

mA HART

: DC , IEC 61010-1. , . ! SITRANS LUT400 , / , .

7ML19985XU81

SITRANS LUT400 (HART)

EL-11

SITRANS LUT400 AC DC.

AC
INPUT POWER

DC
INPUT POWER

12

13

14

12 10 - 32 V

13

14

L1 L2/N 100 - 230 V


AC: 100-230 V AC 15%, 50/60 Hz, 36 VA (10 W)

DC: 10-32 V DC, 10 W

: . 1) , 70 mm (2,75") 1). , : . . AC: 15 A . , , .

2) 3)


SITRANS LUT400 . ( 11 ). (BR2032) . LUT400, . ( ).

1)

, .

EL-12

SITRANS LUT400 (HART)

7ML19985XU81


: . , ( TS-3, ). : SITRANS LUT400 LUT400 ,

TS-3

TRANSDUCER

SHLD

BLK

WHT

()
. ( 2.8.11.Relay Logic ).

BLK

WHT

RELAYS


2, 3 . 1 (NO) (NC). , .

15

16 RLY 1

17

18

19

20

21

RLY 2

RLY 3


C (NO NC) ( 1), 1A 250 V AC, , 3A 30 V DC A (NO) ( 2,3), 5A 250 V AC, , 3A 30 V DC

7ML19985XU81

SITRANS LUT400 (HART)

EL-13

HART mA (HART)
mA OUT/HART

HART ( LUT400), 22 23.

22

23

24

HART ( ), 23 24.

ACT

PAS

, mA (2.5.Current Output) .


36.

SITRANS LUT400

: , (, , ). , ( ). , , ( ). : ( LUI 17) ( : 17)

SITRANS LUT400
. SITRANS LUT400 Measurement () . Siemens LUI, . , (, , , ). . ( , 2.14.Date and Time. ).

EL-14

SITRANS LUT400 (HART)

7ML19985XU81

LCD
Measurement (): 1)
1 10 2 3 4 5
M

SITRANS LUT400 LEVEL

1/2

18.91
1

6
2

5 SAVING PARAMETERS
9 8 7

1 10 11 12 2 3 4

SITRANS LUT400 FLOW HEAD


1 2 3

500313..
60.000

1/2
L/S M

5 6

DAILY TOTAL,...

999999.99

5 SAVING PARAMETERS
9 8 7

1 2 (, , , , ) 3 [ (PV)=1 2, (SV)=2 2] 4 1) PV SV 5 6 7 () () 8 9 10 () : , , , , 11 = = 12 :

0 LOE

8 9

Program Navigation ()
TRANSDUCER 2.1.6 / NO TRANSDUCER , XRS-5 . XPS-10 XPS-15 XCT-8 XCT-12 : XPS-30 . . , . .

1)

7ML19985XU81

SITRANS LUT400 (HART)

EL-15

Parameter ()

Edit ()

SENSOR
UNITS SENSOR MODE SENSOR MODE SECONDARY

2.1.6

TRANSDUCER
NO TRANSDUCER XPS-10 XPS-15 XCT-8 XCT-12 XPS-30

2.1.6

XRS-5

TRANSDUCER

FREQUENCY

XRS-5

PROGRAM
, Program 1. , a) b) c) d) , , .


. . . . LCD . , , (2.2.1.). , . Edit (). .
LOW CALIB. PT. 2.2.1

Edit ().


a) b) c)

ESC

+ _ _ _ _ _ _ 2. 0 M

OK

d)

. e) () , .

f) g)

, .

, ESC. . ,

, OK. h)

. LCD . .

EL-16

SITRANS LUT400 (HART)

7ML19985XU81


SITRANS LUT400 , ( ) , LUI . SIMATIC PDM. SIMATIC PDM 34.

:
1): SIMATIC PDM (HART) AMS (HART) FC375/475 (HART) FDTs (HART) ( ) Web. www.siemens.com/sitransLUT400 > Support () > Application Guides ( ).

LUI
1) 2) Program. (1.), (1.1.) , , : QS Level ( ) (1.1.1.), QS Volume ( ) (1.1.2.) QS Flow ( ) (1.1.3.). [ QS Flow LUI LUT430 ( ) LUT440 (OCM)]. Finish () , Program () Measurement ().

3)

: Finish () . : ' Parameter () . ( ). , .

1. 1.1. 1.1.1. QS Level ( )


. a) b) c)
1)

PROGRAM 1: MAIN MENU ( ). 1.1.1.

QS Level.

SITRANS LUT400 (HART)

7ML19985XU81

SITRANS LUT400 (HART)

EL-17

d)

, Edit (): .

QUICK START LEVEL


TRANSDUCER XRS-5
BACK CANCEL NEXT EDIT

QUICK START LEVEL


TRANSDUCER

e)

, .

NO TRANSDUCER XPS-10 XPS-15 XCT-8 XCT-12

XRS-5

f)

QS Level . CANCEL (), START ()

Siemens . NO TRANSDUCER, XRS-5, XPS-10, XPS-15, XCT-8, XCT-12, XPS30, XPS-40, XLT-30, XLT-60, STH : NO TRANSDUCER ( )

( mA) .

LEVEL
()

( ) ( ) ( )

SPACE
( )

DISTANCE
()

EL-18

SITRANS LUT400 (HART)

7ML19985XU81

( ) . OTHER (), , LEVEL (), SPACE ( ) DISTANCE (). LEVEL, SPACE DISTANCE QS Level.

OTHER ()

. TRANSDUCER, FIXED TEMPERATURE, EXTERNAL TS-3, AVERAGE OF SENSORS : TRANSDUCER () , . : -100,0 +150,0 C : +20,0 C

FIXED TEMPERATURE ( ) . . M, CM, MM, FT, IN : M

: , (m). : . : 0,000 60,000 : 0,000

: . : 0,000 60,000 : 60,000

7ML19985XU81

SITRANS LUT400 (HART)

EL-19

. : , Fill Rate per Minute (2.3.1.), Empty Rate per Minute (2.3.2.) Damping Filter (2.3.3.) . m/min. SLOW () (0,1 M/MIN) MEDIUM () (1,0 M/MIN) FAST () (10 M/MIN) : SLOW (0,1 M/MIN) ( ). , . QS Level , . BACK (), CANCEL (), FINISH () ( , 1.1 . CANCEL, ).

Program (), Measurement (). (). ,

1.1.2. QS Volume ( )
. a) PROGRAM 1: MAIN MENU. 1.1.1. QS Volume.
QUICK START VOLUME
TRANSDUCER XRS-5
BACK CANCEL NEXT EDIT

b) c) d)

, Edit (): . , .

e)

QUICK START VOLUME


TRANSDUCER

NO TRANSDUCER XPS-10 XPS-15 XCT-8 XCT-12

f) EL-20

XRS-5

. SITRANS LUT400 (HART)

7ML19985XU81

QS Volume . CANCEL, START

Siemens . NO TRANSDUCER, XRS-5, XPS-10, XPS-15, XCT-8, XCT-12, XPS30, XPS-40, XLT-30, XLT-60, STH : NO TRANSDUCER TRANSDUCER, FIXED TEMPERATURE, EXTERNAL TS-3, AVGERAGE OF SENSORS : TRANSDUCER , . : -100,0 +150,0 C : +20,0 C

FIXED TEMPERATURE . SITRANS LUT400 . NONE (), . . NONE, LINEAR, CYLINDER, PARABOLIC BOTTOM, HALF SPHERE BOTTOM, FLAT SLOPED BOTTOM, PARABOLIC ENDS, SPHERE, CONICAL BOTTOM, CURVE TABLE, LINEAR TABLE : LINEAR () CURVE TABLE ( ) LINEAR TABLE ( ), . ( 2.6.7.Table 1-8 ). . M, CM, MM, FT, IN : M

: , (m).

7ML19985XU81

SITRANS LUT400 (HART)

EL-21

: . : 0,000 60,000 : 0,000

: . : 0,000 60,000 : 60,000

. : , Fill Rate per Minute (2.3.1.) Empty Rate per Minute (2.3.2.) . m/min. SLOW (0,1 M/MIN) MEDIUM (1,0 M/MIN) FAST (10 M/MIN) : SLOW (0,1 M/MIN) ( ). , . A , , , , .. : 0,000 99,999 : 0,000

L . : 0,000 99,999 : 0,000

. L, USGAL, IMPGAL, CUM, USER DEFINED * : L

* USER DEFINED ( ), . ( 2.6.6.User Defined Unit .

EL-22

SITRANS LUT400 (HART)

7ML19985XU81

. . , 8000 , 8000. : 0,0 9999999 : 100,0

QS Volume , . BACK, CANCEL, FINISH ( , 1.1 ). CANCEL, ).

Program (), Measurement (). (). ,

1.1.3. QS Flow ( )
. [ LUT430 ( ) LUT440 (OCM)]. a) b) c) PROGRAM 1: MAIN MENU. Flow. 1.1.1. QS

d)

, Edit (): .

QUICK START FLOW


TRANSDUCER XRS-5
BACK CANCEL NEXT EDIT

e)

, .

QUICK START FLOW


TRANSDUCER

NO TRANSDUCER XPS-10 XPS-15 XCT-8 XCT-12

XRS-5

f)

. QS Flow . CANCEL, START

7ML19985XU81

SITRANS LUT400 (HART)

EL-23

Siemens . NO TRANSDUCER, XRS-5, XPS-10, XPS-15, XCT-8, XCT-12, XPS30, XPS-40, XLT-30, XLT-60, STH : NO TRANSDUCER . TRANSDUCER, FIXED TEMPERATURE, EXTERNAL TS-3, AVGERAGE OF SENSORS : TRANSDUCER . : -100,0 +150,0 C : +20,0 C

FIXED TEMPERATURE . (PMD) . EXPONENTIAL DEVICES, RECTANGULAR FLUME BS-3680, ROUND NOSE HORIZONTAL CR. BS-3680, TRAPEZOIDAL FLUME BS-3680, U-FLUME BS-3680, FINITE CREST WEIR BS3680, THIN PLATE RECT. WEIR BS-3680, THIN PLATE V-NOTCH WEIR BS-3680, RECT. WEIR CONTRACTED, ROUND PIPE, PALMER BOWLUS FLUME, H-FLUME, OTHER* : EXPONENTIAL DEVICES ( ) * OTHER () HART ( SIMATIC PDM), OFF UNIVERSAL HEAD VS. FLOW. , PMD (OFF) (UNIVERSAL HEAD VS. FLOW) HART (SIMATIC PDM, AMS, FC375/475). . . ABSOLUTE (), RATIOMETRIC () : ABSOLUTE

. M, CM, MM, FT, IN : M

: , (m).

EL-24

SITRANS LUT400 (HART)

7ML19985XU81

: . : 0,000 60,000 : 0,000

: . : 0,000 60,000 : 60,000

. : , Fill Rate per Minute (2.3.1.) Empty Rate per Minute (2.3.2.) . m/min. SLOW (0,1 M/MIN) MEDIUM (1,0 M/MIN) FAST (10 M/MIN) : SLOW (0,1 M/MIN) ( ). , . : : PMD . ( ).

(PMD = EXPONENTIAL DEVICES ( )) . : -999,000 9999,000 : 1,550

7ML19985XU81

SITRANS LUT400 (HART)

EL-25

K (PMD = EXPONENTIAL DEVICES ( )) . : -999,000 9999,000 : 1,000

V (PMD = THIN PLATE V-NOTCH WEIR ( ' )) V . (PMD = TRAPEZOIDAL FLUME ( ) ROUND PIPE ( )) . : -999,000 9999,000 : 0,000 : 25,000 95,000 : 25,000

(PMD = ROUND PIPE ( )) . : -999,000 9999,000 : 0,000

EL-26

SITRANS LUT400 (HART)

7ML19985XU81

PMD

PMD, Exponential Devices ( ) Other (), . PMD, PMD. ( )


APPROACH WIDTH B ( B) (2.15.4.5. OCM Dimension 1) THROAT WIDTH B ( B) (2.15.4.6. OCM Dimension 2) HUMP HEIGHT P ( P) (2.15.4.7. OCM Dimension 3) THROAT LENGTH L ( L) (2.15.4.8. OCM Dimension 4)

PMD

Rectangular Flume ( ) BS-3680

Round Nose Horizontal Crest Weir ( ) BS-3680


CREST WIDTH B ( B) (2.15.4.5. OCM Dimension 1) CREST HEIGHT P ( P) (2.15.4.6. OCM Dimension 2) CREST LENGTH L ( L) (2.15.4.7. OCM Dimension 3)

Trapezoidal Flume ( ) BS-3680


APPROACH WIDTH B ( B) (2.15.4.5. OCM Dimension 1) THROAT WIDTH B ( B) (2.15.4.6. OCM Dimension 2) HUMP HEIGHT P ( P) (2.15.4.7. OCM Dimension 3) THROAT LENGTH L ( L) (2.15.4.8. OCM Dimension 4)

U-Flume ( U) BS-3680
APPROACH DIAMETER DA ( DA) (2.15.4.5. OCM Dimension 1) THROAT DIAMETER D ( D) (2.15.4.6. OCM Dimension 2) HUMP HEIGHT P ( P) (2.15.4.7. OCM Dimension 3) THROAT LENGTH L ( L) (2.15.4.8. OCM Dimension 4)

Finite Crest Weir ( ) BS-3680


CREST WIDTH B ( B) (2.15.4.5. OCM Dimension 1) CREST HEIGHT P ( P) (2.15.4.6. OCM Dimension 2) CREST LENGTH L ( L) (2.15.4.7. OCM Dimension 3)

Thin Plate Rectangular Weir ( ' ) BS3680


APPROACH WIDTH B ( B) (2.15.4.5. OCM Dimension 1) CREST WIDTH B ( B) (2.15.4.6. OCM Dimension 2) CREST HEIGHT P ( P) (2.15.4.7. OCM Dimension 3)

Rectangular Weir Contracted ( )


CREST WIDTH B ( B) (2.15.4.5. OCM Dimension 1) ( )

7ML19985XU81

SITRANS LUT400 (HART)

EL-27

PMD ()

( )
PIPE INSIDE DIAMETER D ( D) (2.15.4.5. OCM Dimension 1)

Round Pipe ( )

Palmer Bowlus Flume ( Palmer Bowlus)


MAXIMUM FLUME WIDTH HMAX ( HMAX) (2.15.4.5. OCM Dimension 1)

H-Flume ( H)
MAXIMUM LISTED HEAD HMAX ( HMAX) (2.15.4.5. OCM Dimension 1)

PMD. : 0,000 60,000 : 60,000

( ) ( ). : -60,000 60,000 : 0,000

. L/S, L/MIN, CUFT/S, CUFT/D, GAL/MIN, GAL/D, IMPGAL/MIN, IMPGAL/D, CUM/H, CUM/D, USER DEFINED * : L/S * USER DEFINED, . ( 2.6.6.User Defined Unit . 20 mA . : -999 9999999 : 100

. NO DIGITS, 1 DIGIT, 2 DIGITS, 3 DIGITS : NO DIGITS ( )

. : 0,000 60,000 : 0,000

EL-28

SITRANS LUT400 (HART)

7ML19985XU81

QS Flow , . BACK, CANCEL, FINISH ( , 1.1.1 ). CANCEL, ).

Program (), Measurement (). : 2.15.2.Auto Zero Head . . (). ,

1.2.
, . . a) b) c) d) PROGRAM 1: MAIN MENU. 1.1. Pump Control.

, . Edit ():

e)

f)

- Pump Control ( ) . CANCEL, START

NONE, 2 : NONE ()

NONE, .

7ML19985XU81

SITRANS LUT400 (HART)

EL-29

1 1. RELAY 2, RELAY 3 : RELAY 2 ( 2)

2 . 2 1 . ( ) 1 = RELAY 2, 2 = RELAY 3 1 = RELAY 3, 2 = RELAY 2

. LUT420: ALTERNATE DUTY ASSIST, ALTERNATE DUTY BACKUP LUT430: ALTERNATE DUTY ASSIST, ALTERNATE DUTY BACKUP, SERVICE RATIO DUTY ASSIST, SERVICE RATIO DUTY BACKUP, FIXED DUTY ASSIST, FIXED DUTY BACKUP OCM LUT440: ALTERNATE DUTY ASSIST, ALTERNATE DUTY BACKUP, SERVICE RATIO DUTY ASSIST, SERVICE RATIO DUTY BACKUP, FIXED DUTY ASSIST, FIXED DUTY BACKUP ( ): ALTERNATE DUTY ASSIST 1 (RUN time ratio) . : 0 255 : 1

. 2 (RUN time ratio) .

: 0 255 : 1

EL-30

SITRANS LUT400 (HART)

7ML19985XU81

2 2, . : 0 999999 : 0

. ( ). 3 3, . : 0 999999 : 0

. ( ). 1 1 2.1.1.Units. : 0,000 99999,000 : 0,000

2 2 2.1.1.Units. : 0,000 99999,000 : 0,000

1 1 2.1.1.Units. : 0,000 99999,000 : 0,000

2 2 2.1.1.Units. : 0,000 99999,000

: 0,000

Pump Control , . BACK, CANCEL, FINISH ( , ). CANCEL, ).

Program (), (). ,

Measurement ().

7ML19985XU81

SITRANS LUT400 (HART)

EL-31

SITRANS LUT400: HART

SITRANS LUT400 HART HART Communicator 375/475 Emerson ( ) . SIMATIC Process Device Manager (PDM) Siemens. Working with HART, Web . : www.siemens.com/sitransLUT400 Support () Application Guides ( ).

SITRANS LUT400 : PC SIMATIC PDM PC AMS Device Manager Emerson PC Web PC Field Device Tool (FDT) Field Communicator 375/475 (FC375/FC475). SITRANS LUT400. SITRANS LUT400 (HART).

SIMATIC PDM
SIMATIC PDM SITRANS LUT400 . www.siemens.com/simatic-pdm. SIMATIC PDM, Service Pack (SP) hot fix (HF). : http://support.automation.siemens.com/WW/ llisapi.dll?func=cslib.csinfo&lang=en&objiD=10806857&subtype=133100

(EDD)
EDD Device Catalog ( ), Sensors/Level/ Echo/Siemens AG/SITRANS LUT400. ( EDD ). EDD, major and minor numbers EDD (.. : 1.00.00-02).
EDD . PDM, SITRANS LUT400 HART > Identification () > Device ().

EDD

SIMATIC PDM Service Pack (SP) hot fix (HF).

EDD
www.siemens.com/sitransLUT400 > Support () > Software Downloads ( ) EDD Web. . SIMATIC PDM Manager Device Catalog, EDD.

EL-32

SITRANS LUT400 (HART)

7ML19985XU81


1) 2) EDD , , ( EDD ). SITRANS LUT400 .

SIMATIC PDM
:
Apply and Transfer ( ) . BACK () CANCEL () .

Device () SIMATIC PDM (, , , ).

AMS Device Manager


AMS Device Manager SITRANS LUT400 . : http://www.emersonprocess.com/AMS/.

(EDD)
SITRANS LUT400, EDD AMS Device Manager, 10.5.

AMS Device Manager


1) EDD Device Catalog ( ), Sensors/ Level/Echo/Siemens AG/SITRANS LUT400. Web www.siemens.com/sitransLUT400, Support ()>Software Downloads ( ), EDD AMS Device Manager. . AMS Device Manager Add Device Type, EDD . ( HART AMS Device Manager Web , Support ()) Device Connection View ( ), Scan Device ( ) . . . SITRANS LUT400 .

2) 3)

4)

5)

6)

7ML19985XU81

SITRANS LUT400 (HART)

EL-33

FC375/475 (HART)
FC375/FC475 HART Communicator , HART, SITRANS LUT400. FC375/475 AMS Device Manager.

(EDD)
SITRANS LUT400 EDD FC375/FC475.

FC375/FC475
1) Web www.siemens.com/sitransLUT400, Support ()->Software Downloads ( ), EDD FC375/FC475. . EDD, Field Communicator Easy Upgrade Utility ( Field Communicator) Emerson. SITRANS LUT400 .

2) 3) 4)

FDT (Field Device Tool)


FDT , SITRANS LUT400. FDT PACTware Fieldcare. FDT PDM. FDT DTM (Device Type Manager) . Siemens SITRANS DTM EDD SITRANS DTM. 1) SITRANS DTM . : http://www.siemens.com/sitransdtm. Support Software downloads. (HART) EDD SITRANS LUT400 SITRANS DTM. Web : www.siemens.com/sitransLUT400 Support Software Downloads.

2)

FDT
Web , Support.

SITRANS LUT400 .

EL-34

SITRANS LUT400 (HART)

7ML19985XU81

: (/) () , .

1. (/)
SITRANS LUT420 7MLxxxx-xxxxx-xxxx Serial No.: GYZ / B1034567 0518 KCC-REM-S49 Power Rating: 100 230V 15% SITRANSLUT400 50/60 Hz, 36 VA (10 W) Contact Rating: 1A/5A @ 250V , Non-Inductive Operating Temperature: 20C to 50C Enclosure: IP65 / TYPE 4X / NEMA 4X II 3D Ex tc IIIC T70C Dc Ex tc IIIC T70C Dc SIRA 12ATEX9143X IECEx SIR 12.0057X Warning: Possible static hazard. Do not rub or clean on site with a dry cloth. Warning: Do not open when energized or when an explosive dust atmosphere is present. Siemens Milltronics Process Instruments, Peterborough Assembled in Canada with domestic and imported parts SITRANS LUT430 7MLxxxx-xxxxx-xxxx Serial No.: GYZ / B1034567 0518 KCC-REM-S49 Power Rating: 100 230V 15% SITRANSLUT400 50/60 Hz, 36 VA (10 W) Contact Rating: 1A/5A @ 250V , Non-Inductive Operating Temperature: 20C to 50C Enclosure: IP65 / TYPE 4X / NEMA 4X II 3D Ex tc IIIC T70C Dc Ex tc IIIC T70C Dc SIRA 12ATEX9143X IECEx SIR 12.0057X Warning: Possible static hazard. Do not rub or clean on site with a dry cloth. Warning: Do not open when energized or when an explosive dust atmosphere is present. Siemens Milltronics Process Instruments, Peterborough Assembled in Canada with domestic and imported parts SITRANS LUT440 7MLxxxx-xxxxx-xxxx Serial No.: GYZ / B1034567 0518 KCC-REM-S49 Power Rating: 100 230V 15% SITRANSLUT400 50/60 Hz, 36 VA (10 W) Contact Rating: 1A/5A @ 250V , Non-Inductive Operating Temperature: 20C to 50C Enclosure: IP65 / TYPE 4X / NEMA 4X II 3D Ex tc IIIC T70C Dc Ex tc IIIC T70C Dc SIRA 12ATEX9143X IECEx SIR 12.0057X Warning: Possible static hazard. Do not rub or clean on site with a dry cloth. Warning: Do not open when energized or when an explosive dust atmosphere is present. Siemens Milltronics Process Instruments, Peterborough Assembled in Canada with domestic and imported parts

SITRANS LUT420 7MLxxxx-xxxxx-xxxx Serial No.: GYZ / B1034567 0518 KCC-REM-S49 Power Rating: 10 32V , 10W SITRANSLUT400 Contact Rating: 1A/5A @ 250V , Non-Inductive Operating Temperature: 20C to 50C Enclosure: IP65 / TYPE 4X / NEMA 4X

SITRANS LUT430 7MLxxxx-xxxxx-xxxx Serial No.: GYZ / B1034567 0518 KCC-REM-S49 Power Rating: 10 32V , 10W SITRANSLUT400 Contact Rating: 1A/5A @ 250V , Non-Inductive Operating Temperature: 20C to 50C Enclosure: IP65 / TYPE 4X / NEMA 4X

SITRANS LUT440 7MLxxxx-xxxxx-xxxx Serial No.: GYZ / B1034567 0518 KCC-REM-S49 Power Rating: 10 32V , 10W SITRANSLUT400 Contact Rating: 1A/5A @ 250V , Non-Inductive Operating Temperature: 20C to 50C Enclosure: IP65 / TYPE 4X / NEMA 4X

II 3D Ex tc IIIC T70C Dc Ex tc IIIC T70C Dc SIRA 12ATEX9143X IECEx SIR 12.0057X Warning: Possible static hazard. Do not rub or clean on site with a dry cloth. Warning: Do not open when energized or when an explosive dust atmosphere is present.

II 3D Ex tc IIIC T70C Dc Ex tc IIIC T70C Dc SIRA 12ATEX9143X IECEx SIR 12.0057X Warning: Possible static hazard. Do not rub or clean on site with a dry cloth. Warning: Do not open when energized or when an explosive dust atmosphere is present.

II 3D Ex tc IIIC T70C Dc Ex tc IIIC T70C Dc SIRA 12ATEX9143X IECEx SIR 12.0057X Warning: Possible static hazard. Do not rub or clean on site with a dry cloth. Warning: Do not open when energized or when an explosive dust atmosphere is present.

Siemens Milltronics Process Instruments, Peterborough Assembled in Canada with domestic and imported parts

Siemens Milltronics Process Instruments, Peterborough Assembled in Canada with domestic and imported parts

Siemens Milltronics Process Instruments, Peterborough Assembled in Canada with domestic and imported parts

ATEX Web www.siemens.com/sitransLUT400. IECEx IECEx. : http://iecex.iec.ch Ex Equipment Certificates of Conformity ( ) IECEx SIR 12.0057X. . ATEX .

7ML19985XU81

SITRANS LUT400 (HART)

EL-35

2. ()
SITRANS LUT420 7MLxxxx-xxxxx-xxxx Serial No.: GYZ / B1034567 Power Rating: 100 230V 15% 50/60 Hz, 36 VA (10 W) Contact Rating: 1A/5A @ 250V , Non-Inductive Operating Temperature: 20C to 50C Enclosure: IP65 / TYPE 4X / NEMA 4X SITRANS LUT430 7MLxxxx-xxxxx-xxxx Serial No.: GYZ / B1034567 Power Rating: 100 230V 15% 50/60 Hz, 36 VA (10 W) Contact Rating: 1A/5A @ 250V , Non-Inductive Operating Temperature: 20C to 50C Enclosure: IP65 / TYPE 4X / NEMA 4X SITRANS LUT440 7MLxxxx-xxxxx-xxxx Serial No.: GYZ / B1034567 Power Rating: 100 230V 15% 50/60 Hz, 36 VA (10 W) Contact Rating: 1A/5A @ 250V , Non-Inductive Operating Temperature: 20C to 50C Enclosure: IP65 / TYPE 4X / NEMA 4X

159134

Class I, Div.2, Gr. A, B, C & D T3C Class II, Div. 2, Gr. F & G Class III Per Drawing A5E03936871

159134

Class I, Div.2, Gr. A, B, C & D T3C Class II, Div. 2, Gr. F & G Class III Per Drawing A5E03936871

159134

Class I, Div.2, Gr. A, B, C & D T3C Class II, Div. 2, Gr. F & G Class III Per Drawing A5E03936871

Siemens Milltronics Process Instruments, Peterborough Assembled in Canada with domestic and imported parts

Siemens Milltronics Process Instruments, Peterborough Assembled in Canada with domestic and imported parts

Siemens Milltronics Process Instruments, Peterborough Assembled in Canada with domestic and imported parts

SITRANS LUT420 7MLxxxx-xxxxx-xxxx Serial No.: GYZ / B1034567 Power Rating: 10 32V , 10W Contact Rating: 1A/5A @ 250V , Non-Inductive Operating Temperature: 20C to 50C Enclosure: IP65 / TYPE 4X / NEMA 4X

SITRANS LUT430 7MLxxxx-xxxxx-xxxx Serial No.: GYZ / B1034567 Power Rating: 10 32V , 10W Contact Rating: 1A/5A @ 250V , Non-Inductive Operating Temperature: 20C to 50C Enclosure: IP65 / TYPE 4X / NEMA 4X

SITRANS LUT440 7MLxxxx-xxxxx-xxxx Serial No.: GYZ / B1034567 Power Rating: 10 32V , 10W Contact Rating: 1A/5A @ 250V , Non-Inductive Operating Temperature: 20C to 50C Enclosure: IP65 / TYPE 4X / NEMA 4X

159134

Class I, Div.2, Gr. A, B, C & D T3C Class II, Div. 2, Gr. F & G Class III Per DWG. A5E03936871

159134

Class I, Div.2, Gr. A, B, C & D T3C Class II, Div. 2, Gr. F & G Class III Per DWG. A5E03936871

159134

Class I, Div.2, Gr. A, B, C & D T3C Class II, Div. 2, Gr. F & G Class III Per DWG. A5E03936871

Siemens Milltronics Process Instruments, Peterborough Assembled in Canada with domestic and imported parts

Siemens Milltronics Process Instruments, Peterborough Assembled in Canada with domestic and imported parts

Siemens Milltronics Process Instruments, Peterborough Assembled in Canada with domestic and imported parts

CSA I, . 2, A5E03936871, Web www.siemens.com/sitransLUT400.

( ATEX 94/9/E, II, 1/0/6)


SIRA 12ATEX9143X: 1) 2) 3) 4) , . 3D. 22. T70 C ( +50 C). . 5) - 20 +50 C. 6) 94/9/E II, 3D, IP . 7) , (T70 C). 8) ( 94/9/E, II, 1.5). 9) (EN 60079-14 EN 60079-17 ). 10) , (.. 60079-19 ).

EL-36

SITRANS LUT400 (HART)

7ML19985XU81

11)

, . 12) .


X : 94/9/E II, 3D, IP . , (T70 C). , , . ( ), . , . , , . LUT400 (LUT420, LUT430, LUT440), , 94/9/E II, 3D IP65 .

7ML19985XU81

SITRANS LUT400 (HART)

EL-37

EL-38

SITRANS LUT400 (HART)

7ML19985XU81

SITRANS LUT400 (HART) Gua rpida


Esta gua expone las principales caractersticas y funciones del dispositivo SITRANS LUT400 (HART 1)). Es altamente recomendable conseguir el manual completo para poder sacar todo el partido al dispositivo. El manual del dispositivo SITRANS LUT400 (HART) con su anexo sobre comunicaciones se puede descargar en la pgina de productos de nuestro sitio web: www.siemens.com/sitransLUT400. El manual impreso se lo puede proporcionar su representante local de Siemens. Para mayor informacin sobre el contenido de esta gua, dirjase a: Siemens AG Siemens Milltronics Process Instruments 1954 Technology Drive, P .O. Box 4225 Peterborough, Ontario, Canad, K9J 7B1 E-mail: techpubs.smpi@siemens.com

Copyright Siemens AG 2012. Todos los derechos reservados


Recomendamos a nuestros usuarios la compra de los manuales impresos originales o la consulta de las versiones electrnicas diseadas y creadas por Siemens Milltronics Process Instruments. Siemens Milltronics Process Instruments no se considerar responsable del contenido de una parte o de la totalidad de las reproducciones de las versiones impresas o electrnicas.

Clusula de exencin de responsabilidad


Nosotros hacemos todo lo necesario para garantizar la conformidad del contenido de esta gua con el equipo proporcionado. Sin embargo, esta informacin queda sujeta a cambios. Por consiguiente, no podemos garantizar su plena conformidad. Examinamos y corregimos regularmente el contenido de esta gua y nos esforzamos por proporcionar publicaciones cada vez ms completas. No dude en ponerse en contacto con nosotros si tiene preguntas o comentarios. Las especificaciones estn sujetas a cambios.

MILLTRONICS es una marca registrada de Siemens Milltronics Process Instruments.

Servicio tcnico
Atencin las 24 horas del da. Para ver la direccin, el nmero de telfono y el nmero de fax de las oficinas locales de Siemens Automation, vaya a: www.siemens.com/automation/partner: Haga clic en la seccin Contact (Contacto) para ver el grupo de productos [por ejemplo: Automation Technology (Tecnologa de automatizacin)]. Vuelva a hacer clic en el cuadro desplegable de grupos de productos para ver las familias que le interesen. Para medicin de nivel: +Automation Technology > +Sensor Systems (Sistemas de deteccin) >+Process Instrumentation (Instrumentacin de procesos) > +Level Measurement (Medicin de nivel) > +Continuous (Procesos continuos). Seleccione Ultrasonic (Ultrasnica). Para caudal: +Automation Technology > +Sensor Systems >+Process Instrumentation > +Flow Measurement (medicin de caudal). Seleccione Ultrasonic. Seleccione el pas seguido por la ciudad o regin. Seleccione Technical Support (Servicio tcnico) en Service (Asistencia). Para el servicio tcnico en lnea, vaya a: www.siemens.com/automation/support-request Ponga el nombre del dispositivo (SITRANS LUT400) o el nmero de referencia, haga clic en Search (Buscar), y seleccione el tipo de producto. Haga clic en Next (Siguiente). Ponga una palabra clave que describa el problema que le ocupa. A continuacin puede buscar la documentacin o hacer clic en Next para enviar por correo una descripcin de su problema al servicio tcnico de Siemens. Centro de servicio tcnico de Siemens IA/DT: telfono +49 (0)911 895 7222
HARTes una marca registrada de HART Communication Foundation.

Espaol

1)

7ML19985XU81

SITRANS LUT400 (HART)

Pgina ES -1

Instrucciones de seguridad
Con objeto de garantizar un uso seguro para el usuario, el personal, el producto y los equipos conectados a ste, es imprescindible seguir las notas de advertencia. En los textos de advertencia se destaca el nivel de precaucin necesario para cada intervencin.

El smbolo de ADVERTENCIA es una seal de precaucin del producto y significa que si no se observan las precauciones de seguridad se puede producir la muerte, lesiones corporales graves y/o daos materiales considerables. El smbolo de ADVERTENCIA, en ausencia del correspondiente smbolo de precaucin del producto, significa que si no se observan las precauciones de seguridad se puede producir la muerte, lesiones corporales graves y/o daos materiales considerables. Nota: significa que se trata de informacin importante sobre el producto o sobre esa parte del manual de instrucciones.

Conformidad FCC
Instalaciones en EE.UU. nicamente: Reglas de la Comisin Federal de Comunicaciones (FCC)
ADVERTENCIA: Al efectuar cambios o modificaciones sin autorizacin previa de Siemens Milltronics se puede anular la autorizacin del usuario a utilizar este dispositivo. Notas: Este equipo ha sido probado y se ha determinado que cumple los lmites asignados a los dispositivos digitales de clase A, segn la parte 15 de las normas FCC. Estos lmites estn pensados para proporcionar una proteccin razonable frente a interferencias que podran perjudicar el funcionamiento del dispositivo en un uso comercial. Este equipo genera, usa y puede irradiar energa de radiofrecuencia y, si no se instala y usa segn las indicaciones del manual de instrucciones, podra provocar interferencias en comunicaciones por radio. El uso de este equipo en reas residenciales podra provocar interferencias a las radiocomunicaciones, en cuyo caso se exigir al usuario que las remedie bajo su propio coste y responsabilidad.

Espaol
Pgina ES-2

SITRANS LUT400 (HART)

7ML19985XU81

SITRANS LUT400
ADVERTENCIA: SITRANS LUT400 slo se puede usar en la forma que se expone en esta gua. De lo contrario podra resultar daada la proteccin del equipo. Nota: Este producto est destinado al uso en zonas industriales. El funcionamiento de este equipo en zonas residenciales puede producir interferencias en las comunicaciones por radiofrecuencia. Los controladores ultrasnicos de largo alcance de la serie Siemens SITRANS LUT400 son compactos y de un solo punto, y miden el nivel en lquidos, lechadas y slidos, adems de monitorizar canales abiertos con gran precisin. La serie es compatible con toda la lnea de transductores EchoMax, lo que significa un rango de funcionamiento de 0,3 a 60 metros (en funcin del transductor). SITRANS LUT400 va acoplado a una interfaz de usuario local retroiluminada (LUI) en la que aparecen los mens de programacin. Se trata de una interfaz intuitiva conectada a red y una serie de asistentes para la instalacin "plug and play". SITRANS LUT400 es compatible con el protocolo HART y con varios paquetes de comunicaciones remotas [SIMATIC PDM, AMS Device Manager, Field Communicator 375/475 y Field Device Tools (FDTs)]. Incorpora adems nuestra ltima generacin de Sonic Intelligence, reforzando an ms nuestro liderazgo en medicin con una facilidad de uso cada vez mayor.

Especificaciones
Alimentacin
Versin AC

1)

En el manual de instrucciones de SITRANS LUT400 (HART) ver una lista completa.

Versin DC
1)

100-230 V AC 15%, 50 / 60 Hz, 36 VA (10W) Fusible: 5 x 20 mm, accin lenta, 0,25A, 250V

10-32 V DC, 10W 1 Fusible: 5 x 20 mm, accin lenta, 1,6A, 125V

El valor de consumo corresponde al pico mximo.

Condiciones ambientales
Altitud
2.000 m mx.

Temperatura ambiente
-20 a +50 C (-4 a +122 F)

Categora de la instalacin
II

Grado de contaminacin
4

Humedad relativa
Apto para uso a la intemperie (slo con caja IP65 / modelo 4X / NEMA 4X)

Homologaciones
Notas: En la etiqueta del dispositivo aparecen sus homologaciones. Asle los tubos cumpliendo los requisitos de la calificacin IP o NEMA. General Peligroso No inflamable (Canad) A prueba de ignicin de polvo (Internacional/Europa) CSAUS/C, CE, FM, homologacin UL, C-TICK CSA clase I, div. 2, grupos A, B, C, D; clase II, div 2, grupos F, G; clase III ATEX II 3D Ex tc IIIC T70 C IP65 IECEx SIR 12.0057X SIRA 12ATEX9143X

Espaol

7ML19985XU81

SITRANS LUT400 (HART)

Pgina ES -3

Instalacin
Notas: La instalacin ser efectuada por personal cualificado, respetando los reglamentos vigentes en el mbito local. Este producto es susceptible de choques electrostticos. Conctelo correctamente a tierra. Asle todas las conexiones teniendo en cuenta que la tensin mnima ser de 250V. Los conectores del transductor estn bajo tensiones potencialmente peligrosas durante su funcionamiento. En los conectores de entrada de corriente continua debern enchufarse cables con aislamiento elctrico entre la entrada y la salida, cumpliendo as los requisitos de seguridad de la norma IEC 61010-1. Los conectores de contacto a rels se utilizan con equipos que no tienen partes activas accesibles y unas conexiones con un aislamiento adecuado para 250 V como mnimo. El voltaje operativo mximo permisible entre los rels adyacentes ser de 250 V. La caja no metlica no tiene toma a tierra entre las conexiones de los tubos. Utilice bornas y puentes conectables a tierra.

Instrucciones de montaje
Las instrucciones de montaje difieren segn el tipo de instalacin del dispositivo: mural, en tubo, en riel DIN o con control remoto. Siga las instrucciones correspondientes a su dispositivo. Nota: Cuando meta los cables en un tubo, siga las instrucciones de la seccin Tendido de cables de la pgina 9 antes de montar el SITRANS LUT400.

Montaje en pared o en panel


Todas las versiones de SITRANS LUT400 se entregan con una chapa base de montaje. El modelo de panel incorpora una pantalla de interfaz de usuario local y una tapa transparente.
chapa base de montaje

caja con tapa opcional para la pantalla enganches (2 unidades) enganches (2 unidades)

Espaol

Nota: Las sujeciones murales no estn incluidas.

Pgina ES-4

SITRANS LUT400 (HART)

7ML19985XU81

chapa base de montaje

caja con tapa transparente enganches (2 unidades) enganches (2 unidades)

Nota: Las sujeciones murales no estn incluidas.

vista posterior de la caja

orificios de montaje

Montaje mural de la caja


1) 2) 3) Marque y perfore cuatro orificios correspondientes a los cuatro tornillos (provistos por el cliente). Apritelos con un destornillador. Alinee los orificios de montaje de la parte posterior del dispositivo con los enganches de la chapa base de montaje. Empuje el LUT400 alinendolo bien contra la chapa base y deslcelo hacia abajo hasta que encaje.

orificios de los tornillos en la chapa base

A tener en cuenta:
Tamao recomendado de las sujeciones: Tornillo M8 o 5/16" con arandela de 17 mm como mximo o 5/8" de dimetro exterior Montaje recomendado: directamente en la pared. En caso de usar otra superficie de montaje, TIENE que ser capaz de soportar el cudruple del peso del dispositivo.

Espaol

7ML19985XU81

SITRANS LUT400 (HART)

Pgina ES -5

Tapa montada remotamente

tapa de pantalla remota

marco

incluidas 4 sujeciones para montaje en panel cable alargador SITRANS LUT400

Montaje de la tapa remota


Nota: La tapa remota se puede montar hasta a 5 m del dispositivo mediante dos cables alargadores opcionales (cada uno de 2,5 m). Consulte el manual de instrucciones. 1) Utilice la plantilla para recortar en el panel el orificio de la tapa de la pantalla remota. Coloque el marco dentro de la tapa, alineando los orificios de montaje. Alinee la parte posterior de la tapa con el recorte hecho en el panel. Marque y taladre los cuatro orificios de los tornillos (incluidos) en la superficie de montaje. Apritelos con un destornillador y una llave. Nota: Par de apriete recomendado de los tornillos para un buen sellado: 1,1 N m 10 in-lbs

2)

A tener en cuenta:
Montaje recomendado: en panel hasta 5 m de distancia del dispositivo. En caso de usar otra superficie de montaje, TIENE que ser capaz de soportar el cudruple del peso del dispositivo. Nota: Sujeciones incluidas: tornillo M5, arandela de sello, arandela plana y tuerca M5. Estas sujeciones son necesarias para garantizar la proteccin IP54 de la tapa remota.

Espaol

Pgina ES-6

SITRANS LUT400 (HART)

7ML19985XU81

Montaje en tubo

chapa base de montaje abrazadera

U roscada

caja con tapa opcional para la pantalla

tubo Nota: Las sujeciones para tubo no estn incluidas.

Montaje de la caja
1) 2) 3) Sujete la chapa base en el tubo mediante las U roscadas y las abrazaderas adecuadas para el dimetro del tubo (provistas por el cliente). Apriete los tornillos con una llave. No los apriete demasiado para evitar que la chapa se tuerza o se doble, lo que podra impedir la sujecin del LUT400 en la chapa base. Sujete el dispositivo en la chapa base (como se describe en el paso 3 de Montaje mural de la caja en la pgina 5).

A tener en cuenta:
Montaje recomendado: directamente en un tubo horizontal o vertical. En caso de usar otra superficie de montaje, TIENE que ser capaz de soportar el cudruple del peso del dispositivo. Dimensiones recomendadas del tubo: mximo: 3", mnimo: 3/4" Tamao recomendado de las sujeciones: U roscadas: mximo: para tubo de 3" con rosca M8 o 3/8" mnimo: para tubo de 3/4" con rosca M6 o 1/4" Tuercas hexagonales: M6 o 1/4" a M8 o 3/8" Arandela: mximo: 16 mm o 13/16" de dimetro externo.

Espaol

7ML19985XU81

SITRANS LUT400 (HART)

Pgina ES -7

Montaje sobre riel DIN


vista posterior de la caja

riel DIN

enganches (2)

ranura

gua del riel DIN

Montaje de la caja
1) 2) 3) Incline la parte superior de la caja y colquela ligeramente por encima de la parte superior del riel DIN. Baje la caja hasta que los enganches de la parte posterior de la caja encajen en el riel. Empuje el dispositivo alinendolo bien contra el riel para enganchar la gua, que ser la que sujete definitivamente la caja en el riel.

A tener en cuenta:
Montaje recomendado: directamente en riel DIN horizontal. Dimensiones necesarias del riel DIN: TH 35-7,5 o TH 35-15 por cada IEC 60715 estndar. El riel DIN TIENE que ser capaz de soportar el cudruple del peso del SITRANS LUT400.

Desinstalacin de la caja
1) Mirando el dispositivo de frente, coloque un destornillador en la ranura de la parte inferior del riel y haga palanca hacia abajo. De esta forma se soltarn los enganches de la parte inferior del riel. Manteniendo presionada la gua, levante la caja para soltar los enganches del riel.

2)

Espaol
Pgina ES-8

SITRANS LUT400 (HART)

7ML19985XU81

Preparacin para la introduccin de los cables


Los cables se pueden llevar por un tubo o introducirlos en la caja a travs de prensaestopas. Siga los pasos 1 a 5 de ms abajo para dejar al descubierto los orificios de entrada de los cables y, a continuacin, proceda segn vaya a usar tubo de cables o prensaestopas. 1) 2) 3) 4) 5) Cierre bien la tapa de la caja y apriete bien los tornillos de sujecin. Coloque la punta del destornillador en la ranura de la circunferencia exterior de la tapa del orificio (consulte la ilustracin de ms abajo). Golpee con la palma de la mano el extremo del destornillador para levantar la tapa del orificio. Afloje los tornillos y quite la tapa de la caja. Quite las tapas de plstico de los orificios. Tenga cuidado de no daar la electrnica con electricidad esttica o con las herramientas que use para quitar las tapas de plstico.
SITRANS LUT400

orificios de entrada del cable

tapa de orificio (3 unidades)

ranura para el destornillador (3 unidades)

Notas: Cuando haya quitado las tapas de los orificios, el dimetro de stos ser de 21,4 a 21,6 mm. Los prensaestopas de cable M20 (20 mm de dimetro) y el tubo NPT de 1/2" (de 21,3 mm de dimetro) pasan sin problemas por estos orificios. Elija con cuidado lo que utilice para sellarlos. Se recomienda el uso de juntas planas (en lugar de tricas). Si se usan otros prensaestopas, el cliente ser responsable de garantizar la calificacin IP65 de los orificios de entrada.

Cable pasado por un tubo


(continuacin de los pasos 1 a 5 de ms arriba) 6) Despus de haber preparado la introduccin de los cables tal como se explica en los pasos 1 a 5, use un tubo cuyo tamao est homologado para aplicaciones estancas y ajstelo en la caja. (El tamao es 1/2" NPT.) 7) Vuelva a colocar la tapa de la caja y apriete los tornillos.

Cable expuesto e introducido en los prensaestopas


(continuacin de los pasos 1 a 5 de ms arriba) 6) Despus de haber preparado la introduccin de los cables tal como se explica en los pasos 1 a 5, desatornille los prensaestopas y sujtelos firmemente en la caja. 7) Atornille los cables mediante los prensaestopas. El cable de potencia deber estar separado de los cables de seal. Enchufe los cables en sus respectivos conectores. 8) Apriete los prensaestopas para crear un buen aislamiento. 9) Vuelva a colocar la tapa de la caja y apriete los tornillos.

Espaol

7ML19985XU81

SITRANS LUT400 (HART)

Pgina ES -9

Conexin
Notas de seguridad para la conexin
En el manual de instrucciones encontrar todas las indicaciones para realizar las conexiones. Compruebe si todos los componentes del sistema estn instalados segn las instrucciones. Conecte todos los cables apantallados en los conectores apantallados del LUT400 (sealados con el smbolo ). Para evitar diferencias de potencial a tierra, los cables apantallados debern estar correctamente derivados a tierra. Coloque los hilos conductores dentro de cables apantallados lo ms cortos posible para reducir distorsiones en la lnea provocadas por dispersiones y captaciones de ruido.

Conexin de SITRANS LUT400


ADVERTENCIAS:

Mire la etiqueta del dispositivo para comprobar su calificacin. Asle los tubos cumpliendo los requisitos de la calificacin IP o NEMA.
Notas:
Puede que tenga que separar los cables y los tubos para cumplir las prcticas estndar de conexiones de dispositivos o los cdigos de electricidad.

Para llegar hasta la placa de puertos: 1) 2) 3) Afloje 1/4 de vuelta los tornillos de ajuste. Levante la tapa hacia la izquierda, sobre sus bisagras. La tapa se puede dejar abierta con sus bisagras, o bien se puede sacar de stas y retirarla a un lado para llegar hasta la placa de puertos.

4) 5) 6)

Realice todas las conexiones segn se explica en las siguientes instrucciones. Cuando haya terminado de conectar los cables, vuelva a colocar la tapa. Apriete los tornillos de ajuste.

Espaol
Pgina ES-10

SITRANS LUT400 (HART)

7ML19985XU81

Placa de puertos
En la placa de puertos del LUT400 se pueden conectar simultneamente todas las entradas y salidas. Para facilitar las conexiones se pueden quitar las regletas. ATENCIN: Asegrese de poner las regletas en su conector correspondiente cuando las vuelva a instalar. De lo contrario se podra daar el dispositivo o los equipos externos que tenga conectados. Nota: Par recomendado en los tornillos de sujecin de conectores: 0,56 a 0,79 N m 5 a 7 in-lbs No apriete los tornillos en exceso.

alimentacin
TS-3 TRANSDUCER DISCRETE INPUTS INPUT POWER

alimentacin elctrica

8 1

9 2

10

11 0.25 A, 250 V SLOW BLOW

12 L1

13 L2/N 100 - 230 V

14

DC SYNC

SHLD

BLK

WHT

BLK

WHT

conectores
mA OUT/HART BATTERY BR2032

batera

RELAYS

DISPLAY 15 16 RLY 1 17 18 19 20 21 22 23 24

RLY 2

RLY 3 ACT

puerto USB
SERVICE

PAS

cable de monitor

conexin mA HART

Alimentacin
ADVERTENCIAS: En los conectores de entrada de corriente continua debern enchufarse cables con aislamiento elctrico entre la entrada y la salida, cumpliendo as los requisitos de seguridad de la norma IEC 61010-1. Asle todas las conexiones teniendo en cuenta la tensin nominal. Importante Antes de enchufar SITRANS LUT400 a la corriente por primera vez, compruebe si estn desactivados todos los equipos de alarma o control hasta ver si el sistema funciona correctamente.

Espaol

7ML19985XU81

SITRANS LUT400 (HART)

Pgina ES -11

SITRANS LUT400 est disponible en versin AC o DC.

Alimentacin AC
INPUT POWER

Alimentacin DC
INPUT POWER

12

13

14

12 10 - 32 V

13

14

L1 L2/N 100 - 230 V

AC: 100-230 V AC 15%, 50 / 60 Hz, 36 VA (10W)

DC: 10-32 V DC, 10W

Nota: Asegrese de conectar el dispositivo a una toma a tierra fiable. 1) 2) 3) Para conectarlo a la corriente, pele el extremo del cable unos 70 mm (2,75") y enrosque los hilos a travs del prensaestopas 1). Conecte los cables en los conectores tal como se indica: la polaridad aparece sealada en cada uno de los conectores. Ponga a tierra el dispositivo observando las disposiciones locales. Notas para conexiones de corriente alterna: Si se instala en un edificio, el equipo tiene que estar protegido por un fusible de 15 A o por un disyuntor. Adems, en este caso hay que instalar al lado del equipo y al alcance del operador un disyuntor o un conmutador marcado como conmutador de desconexin.

Batera
SITRANS LUT400 incluye una batera. (Consulte su ubicacin en la pgina 11 .) La batera (BR2032) tiene una vida til de diez aos y se ve afectada por la temperatura ambiente. Si el dispositivo LUT400 dejase de recibir alimentacin elctrica, la batera mantendra en hora su reloj hasta el retorno de la corriente. (En el manual de instrucciones podr ver cmo cambiarla.)

Espaol
1)Si lleva el cable por un tubo, use concentradores de tamao adecuado y homologados para aplicaciones

estancas.

Pgina ES-12

SITRANS LUT400 (HART)

7ML19985XU81

Transductores
Advertencia: Los conectores de los transductores estn bajo tensiones potencialmente peligrosas durante su funcionamiento. Meta el cable de los transductores en un tubo metlico con toma a tierra, y seprelo de las dems conexiones (salvo la del sensor de temperatura TS-3, si lo hubiera).

Notas: No use cables coaxiales con SITRANS LUT400, ya que su blindaje recibira la alta tensin transmitida No conecte juntos los cables de transductor apantallados y blancos: seprelos en diferentes conectores. Ignore los antiguos manuales de transductores que recomendaban estas prcticas
TS-3 TRANSDUCER

SHLD

BLK

WHT

Rels
Los contactos a rel aparecen en posicin desactivada. Todos los rels pueden configurarse con lgica positiva o negativa (consulte el parmetro 2.8.11.Relay Logic en el manual de instrucciones).
RELAYS

BLK

WHT

Corte de corriente
Los rels 2 y 3 estn normalmente abiertos. El rel 1 se puede conectar normalmente abierto o normalmente cerrado. En caso de corte de corriente elctrica, los rels volvern a su estado normal.

15

16 RLY 1

17

18

19

20

21

RLY 2

RLY 3

Capacidades nominales de los rels


un rel de forma C (NO o NC) (rel 1), 1A a 250 V AC, no inductivo, 3A a 30 V DC dos rels de forma A (NO) (rels 2,3), 5A a 250 V AC, no inductivos, 3A a 30 V DC

Espaol

7ML19985XU81

SITRANS LUT400 (HART)

Pgina ES -13

Conexin HART Salida mA (HART)


mA OUT/HART

En caso de conexin HART ACTIVA (con la alimentacin elctrica que incorpora el LUT400), enchufe los conectores 22 y 23.

22

23

24

En caso de conexin HART PASIVA (con alimentacin elctrica externa), enchufe los conectores 23 y 24.

ACT

PAS

Si desea ms informacin, consulte los parmetros de salida mA (2.5.Current Output) en la seccin del manual de instrucciones que explica los parmetros.

Conexin en reas peligrosas


Vase pgina 35.

Puesta en marcha de SITRANS LUT400


Recomendamos su configuracin siguiendo esta secuencia: Abra el asistente de la aplicacin que vaya a usar (nivel, volumen o caudal). Configure las bombas con el asistente de bombas (si lo hubiera). Configure las alarmas o controles, los totalizadores y los muestreadores, especificando los respectivos parmetros (consulte el manual de instrucciones). Hay dos maneras de acceder a los asistentes: localmente (consulte Asistentes de la interfaz de usuario en la pgina 17) a distancia (consulte Otros asistentes: en la pgina 17)

Activacin de SITRANS LUT400


Encienda el dispositivo. SITRANS LUT400 se inicia automticamente en el modo de medicin. En pantalla aparece una ventana de transicin con el logotipo de Siemens primero y, despus, con la versin actual del fabricante de la interfaz, mientras se procesan las primeras mediciones. La primera vez que se configura el dispositivo aparece la opcin de elegir el idioma (ingls, alemn, francs, espaol o chino). Antes de nada deber poner la fecha y la hora correctas. (Para modificarlas, ponga los datos de 2.14.Date and Time. Consulte los detalles en el manual de instrucciones).

Espaol
Pgina ES-14

SITRANS LUT400 (HART)

7ML19985XU81

Pantalla LCD Visualizacin en modo de medicin Funcionamiento normal 1) Nivel


1 10
SITRANS LUT400 NIVEL

2
1/2

3
M

4 5

18.91
1

5 GUARDAR PARMETROS

Caudal
1 10 11 12
SITRANS LUT400 CAUDAL ALTURA
1 2 3

3
1/2
L/S M

4 5 6

500313..
60.000

TOTALIZADO...

999999.99

5 GUARDAR PARMETROS

1 etiqueta 2 valor medido (nivel, espacio, distancia, volumen, caudal o altura mxima) 3 valor que aparece en pantalla [variable principal (PV)=1 de 2, variable secundaria (SV)=2 de 2] 4 cambiar de indicador1) para PV o SV 5 unidades 6 la grfica de barras indica el nivel 7 la regin secundaria indica los rels configurados (izquierda) y las entradas analgicas (derecha) 8 el rea de texto muestra los mensajes de estado 9 indicador de estado del dispositivo 10 modo del sensor seleccionado (principal): nivel, espacio, distancia, volumen, altura mxima y caudal 11 modo de sensor secundario = altura mxima cuando el principal = flujo 12 valores del totalizador: aparecen alternativamente el totalizador diario y el totalizador contador

Error
0 LOE

8 el rea de texto muestra un cdigo de error junto con el mensaje 9 aparece el icono de servicio tcnico

Visualizacin en modo de programacin Vista de navegacin


Cuando aparece la barra de men TRANSDUCTOR 2.1.6 / nmero de desplazamiento de mens significa que actual nombre del parmetro SIN TXT parmetro actual la lista es demasiado larga y no caben XRS-5 XPS-10 todos en pantalla. barra de XPS-15 elemento XCT-8 El ancho de la barra indica la longitud de elementos actual XCT-12 la lista de mens: cuanto ms ancha barra de XPS-30 desplazamiento sea, menos elementos tendr. de mens La posicin de la barra indica el puesto aproximado del elemento marcado en la lista. Una barra que est a la mitad significar que el artculo marcado est en el centro de la lista. Cuanto ms ancha sea, menos elementos tendr.

Espaol

1)

Pulse

para pasar al otro modo.

7ML19985XU81

SITRANS LUT400 (HART)

Pgina ES -15

Vista de parmetros
men en el que se encuentra el parmetro

Vista de edicin
2.1.6
nmero de parmetro actual

CALIBR. SENSOR
UNIDAD (SENSOR) FUNCIONAMIENTO MODO SENSOR SECUNDARIO

TRANSDUCTOR
SIN TXT XPS-10 XPS-15 XCT-8 XCT-12 XPS-30

2.1.6

XRS-5

TRANSDUCTOR

nombre del parmetro

FRECUENCIA

XRS-5

valor o seleccin de parmetros

Modo PROGRAM (programacin)


Presione los pulsadores manuales para entrar en el modo Program (programacin) y abrir el , , y . nivel 1 de mens. Recorra los elementos del men con

Seleccin de una opcin


a) Vaya al parmetro que desee. b) Pulse la flecha a la DERECHA para abrir el modo de edicin. En pantalla aparece resaltado el elemento seleccionado. c) Vaya a una nueva seleccin. d) Pulse la flecha a la DERECHA para aceptarla. La pantalla LCD vuelve a la vista de parmetros y muestra la nueva seleccin.

Cambio de un valor numrico


a) Vaya al parmetro deseado, por ejemplo, Low Calibration Point (2.2.1.) (punto de calibracin inferior). b) Una vez seleccionado, el valor aparecer en pantalla. c) Pulse la flecha a la DERECHA para abrir el modo de edicin. La posicin del cursor aparece resaltada. d) Use las flechas a la IZQUIERDA y a la DERECHA para llevar el cursor hasta el dgito que desea cambiar. e) Como cada dgito aparece resaltado (seleccionado), use las flechas de ARRIBA ABAJO f) para aumentarlo o reducirlo. y ABAJO para hasta y
PT. CALIBR. MNIMO 2.2.1

ESC

+ _ _ _ _ _ _ 2. 0 M

OK

Si est seleccionada la coma decimal, use la flecha ARRIBA cambiar la posicin.

g) Para salir sin guardar los cambios, pulse varias veces la flecha a la IZQUIERDA

que ESC aparezca resaltado. Pulse de nuevo la flecha a la IZQUIERDA para salir sin guardar los cambios. Si el valor del parmetro es correcto, pulse la flecha a la DERECHA continuamente hasta que aparezca resaltada la tecla Aceptar.

Espaol

h) Pulse la flecha a la DERECHA para aceptar el valor. La pantalla LCD vuelve a la vista de parmetros y muestra la nueva seleccin. Revsela para comprobar si est bien.

Pgina ES-16

SITRANS LUT400 (HART)

7ML19985XU81

Asistentes
Si va a configurar SITRANS LUT400 para aplicaciones de nivel, volumen (depsitos de forma estndar) o caudal, use los Asistentes de la interfaz de usuario descritos ms adelante. En SIMATIC PDM encontrar un asistente para aplicaciones de tanques con formas ms complejas. Vase Asistentes en SIMATIC PDM en la pgina 33.

Otros asistentes:
Existen otros asistentes que usan diversos paquetes de software 1): SIMATIC PDM (HART) AMS (HART) FC375/475 (HART) FDTs (HART) En nuestro sitio web encontrar tablas de configuracin de parmetros (para registrar datos antes de configurar el dispositivo con el asistente). Vaya a www.siemens.com/sitransLUT400 > Support (Servicio tcnico) > Application Guides (Manuales de aplicaciones).

Asistentes de la interfaz de usuario


1) 2) Pulse para entrar en el modo Program (programacin). Seleccione Wizards (Tutoriales) (1.), Quick Start (Asistentes) (1.1.) y el asistente que le interese: QS Level (Nivel QS) (1.1.1.), Volumen QS (1.1.2.), o bien Caudal QS (1.1.3.) [El asistente de caudal slo mostrar en la interfaz los modelos LUT430 (bombas y caudal) y LUT440 (OCM) configurados]. Siga los pasos correspondientes y pulse Finish (Terminar) para guardar los cambios en los parmetros y volver al men Program (programacin). Pulse a continuacin tres veces para volver al modo de medicin. Notas: Los parmetros de los asistentes estn interrelacionados, por lo que los cambios slo se harn efectivos al pulsar Finish en el ltimo paso. No use el asistente para modificar parmetros sueltos: consulte para ello la seccin Parmetros del manual de instrucciones. (Caracterice la aplicacin cuando haya terminado el asistente). Si desea ms informacin sobre parmetros de los asistentes, consulte el manual de instrucciones.

3)

1. Wizards (Tutoriales) 1.1. Quick Start (Asistentes) 1.1.1. QS Level (Nivel QS)
Use este asistente para configurar aplicaciones de nivel sencillas. a) Pulse la flecha a la DERECHA para activar el modo PROGRAM (programacin) y abrir el nivel de mens 1: MEN PRINCIPAL b) Pulse dos veces la flecha a la DERECHA c) Pulse la flecha a la DERECHA para ir al elemento 1.1.1. del men.

Espaol

para abrir QS Level (nivel QS).

1)Consulte el anexo sobre comunicaciones del manual SITRANS LUT400 (HART)

7ML19985XU81

SITRANS LUT400 (HART)

Pgina ES -17

d) Pulse siempre la flecha hacia ABAJO para aceptar los valores predeterminados y pasar directamente al elemento siguiente, o la flecha a la DERECHA para abrir el modo de edicin: en pantalla aparece resaltada la seleccin. e) Desplcese hasta el elemento que le interesa, pulse la flecha a la DERECHA el cambio y pulse la flecha ABAJO continuar. f) para guardar para

PUESTA EN MARCHA RAPIDA - NIVEL


TRANSDUCTOR XRS-5
RETORNO ANULAR SIGUIENTE EDITAR

PUESTA EN MARCHA RAPIDA - NIVEL


TRANSDUCTOR

SIN TXT XPS-10 XPS-15 XCT-8 XCT-12

XRS-5

En cualquier momento puede pulsar la flecha hacia ARRIBA IZQUIERDA para cancelar el asistente.

para volver o a la

Inicio del asistente de QS Level (nivel QS) Muestra el tipo de asistente que se va a ejecutar.

Opciones

CANCEL (cancelar), START (inicio)

Transductor Determina qu transductor Siemens est conectado al dispositivo.

Opciones

NO TRANSDUCER (sin transductor), XRS-5, XPS-10, XPS-15, XCT-8, XCT-12, XPS-30, XPS-40, XLT-30, XLT-60, STH Predeterminado: NO TRANSDUCER

Funcionamiento Establece el tipo de medicin (y la salida mA correspondiente) necesaria para la aplicacin.


punto de referencia del sensor punto de calibracin superior distancia espacio

nivel punto de calibracin inferior rango mximo

Modo
NIVEL

Descripcin
* Altura del material

Punto de referencia
Punto de calibracin inferior (en vaco)

Espaol

ESPACIO DISTANCIA

Punto de calibracin superior Distancia a la superficie del material (lleno) Punto de referencia del sensor (contina en la pgina siguiente)

Pgina ES-18

SITRANS LUT400 (HART)

7ML19985XU81

(viene de la pgina anterior)

Modo

Descripcin
NO lo seleccione. Si el valor Operation (funcionamiento) que aparece es OTHER (otros), el dispositivo se configura como un controlador de nivel, pero anteriormente estuvo en un modo que no era ni LEVEL (nivel), ni SPACE (espacio), ni DISTANCE (distancia). El modo de funcionamiento se tiene que poner en LEVEL, SPACE, o DISTANCE para continuar con el asistente de QS Level (nivel QS).

Punto de referencia

OTHER (otros)

Fuente de temperatura Es la fuente de la lectura de temperatura que se va a usar para poner la velocidad del sonido..

Opciones

TRANSDUCER (transductor), FIXED TEMPERATURE (temperatura fija), EXTERNAL TS-3 (TS-3 externo), AVERAGE OF SENSORS (media de sensores) Predeterminado: TRANSDUCER

Temperatura fija Use esta opcin si no va a utilizar un sistema de monitorizacin de la temperatura.

Valor

Rango: -100,0 a +150,0 C Predeterminado: +20,0 C

Este parmetro slo aparece si est seleccionada la opcin FIXED TEMPERATURE (Temperatura fija) en Fuente de temperatura.

Unidades Unidades de medida del sensor.

Opciones

M, CM, MM, FT, IN Predeterminado: M

Nota: En este ejemplo se supone que todos los valores aparecen en metros (m).

Punto de calibracin superior Distancia del punto de referencia del sensor al punto de calibracin superior: normalmente el nivel del depsito lleno.

Valor

Rango: 0,000 a 60.000 Predeterminado: 0,000

Punto de calibracin inferior Distancia del punto de referencia del sensor al punto de calibracin inferior: normalmente el nivel del depsito vaco.

Espaol

Valor

Rango: 0,000 a 60.000 Predeterminado: 60.000

7ML19985XU81

SITRANS LUT400 (HART)

Pgina ES -19

Velocidad de respuesta Determina la velocidad de reaccin del dispositivo a los cambios de medida en el rango objetivo.
Notas: La velocidad de respuesta slo se puede determinar con el asistente, y cualquier cambio realizado en los parmetros Fill Rate per Minute (2.3.1.), Empty Rate per Minute (2.3.2.) o Damping Filter (2.3.3.) despus de haber terminado ste sobrescribir el valor de la velocidad de respuesta. Velocidad de respuesta siempre aparece en m/minuto.
SLOW (lento) (0,1 M/MIN) MEDIUM (medio) (1,0 M/MIN) FAST (rpido) (10 M/MIN) Predeterminado: SLOW (lento) (0,1 M/MIN)

Opciones

Use un valor ligeramente superior a la velocidad mxima de llenado o vaciado (cualquiera que sea mayor). Con una velocidad baja se consigue mayor precisin, y con una alta se registran mejor las fluctuaciones rpidas de nivel.

Fin del asistente de QS Level (nivel QS) Para que QS funcione correctamente, hay que guardar todos los cambios.

Opciones

BACK (anterior), CANCEL (cancelar), FINISH (terminar) [La pantalla vuelve al men 1.1 Quick Start (asistente) cuando el asistente ha terminado o se ha cancelado. Si se ha seleccionado CANCEL, no quedar registrado ningn cambio].

Para transferir los valores del asistente al dispositivo y volver al men Program (programacin), pulse la flecha hacia ABAJO veces la flecha a la IZQUIERDA (Terminar). Pulse a continuacin tres para volver al modo de medicin.

1.1.2. Volumen QS
Use este asistente para configurar las aplicaciones de volumen que utilicen depsitos de forma estndar. a) Pulse la flecha a la DERECHA para activar el modo PROGRAM (programacin) y abrir el nivel de mens 1: MEN PRINCIPAL b) Pulse dos veces la flecha a la DERECHA c) Pulse la flecha hacia ABAJO (volumen QS). para ir al elemento 1.1.1. del men. para abrir el men QS Volume

y a la DERECHA

d) Pulse siempre la flecha hacia ABAJO para aceptar los valores predeterminados y pasar directamente al elemento siguiente, o pulse la flecha a la DERECHA para abrir el modo de edicin: en pantalla aparece resaltada la seleccin. e) Desplcese hasta el elemento que le interesa, pulse la flecha a la DERECHA para guardar el cambio y pulse la

PUESTA EN MARCHA RAPIDA - VOLU..


TRANSDUCTOR XRS-5
RETORNO ANULAR SIGUIENTE EDITAR

PUESTA EN MARCHA RAPIDA - VOLUM..


TRANSDUCTOR

SIN TXT XPS-10 XPS-15 XCT-8 XCT-12

Espaol

XRS-5

flecha ABAJO f)

para continuar. para

En cualquier momento puede pulsar la flecha hacia ARRIBA para volver o a la IZQUIERDA cancelar el asistente.

Pgina ES-20

SITRANS LUT400 (HART)

7ML19985XU81

Inicio del asistente de QS Volume (volumen QS)


Muestra el tipo de asistente que se va a ejecutar.

Opciones

CANCEL (cancelar), START (inicio)

Transductor Determina qu transductor Siemens est conectado al dispositivo.

Opciones

NO TRANSDUCER (sin transductor), XRS-5, XPS-10, XPS-15, XCT-8, XCT-12, XPS-30, XPS-40, XLT-30, XLT-60, STH Predeterminado: NO TRANSDUCER

Fuente de temperatura Es la fuente de la lectura de temperatura que se va a usar para poner la velocidad del sonido.

Opciones

TRANSDUCER (transductor), FIXED TEMPERATURE (temperatura fija), EXTERNAL TS-3 (TS-3 externo), AVERAGE OF SENSORS (media de sensores) Predeterminado: TRANSDUCER

Temperatura fija Use esta opcin si no va a utilizar un sistema de monitorizacin de la temperatura.

Valor

Rango: -100,0 a +150,0 C Predeterminado: +20,0 C

Este parmetro slo aparece si est seleccionada la opcin FIXED TEMPERATURE (Temperatura fija) en Fuente de temperatura.

Forma del depsito


Define la forma del depsito, en base a la cual SITRANS LUT400 calcula el volumen en lugar del nivel. Si est seleccionado NONE (ninguna), no se realizar ninguna conversin de volumen. Seleccione la forma que coincida con el depsito o tanque que vaya a monitorizar.
NONE (ninguna), LINEAR (lineal), CYLINDER (cilndrica), PARABOLIC BOTTOM (base parablica), HALF SPHERE BOTTOM (base semiesfrica), FLAT SLOPED BOTTOM (base con inclinacin plana), PARABOLIC ENDS (extremos parablicos), SPHERE (esfrico), CONICAL BOTTOM (base cnica), CURVE TABLE (grfica curva), LINEAR TABLE (grfica lineal) Predeterminado: LINEAR

Opciones

Si selecciona CURVE TABLE (grfica curva) o LINEAR TABLE (grfica lineal), ponga los valores de interrupcin de nivel y de volumen cuando haya terminado el asistente. (Consulte la tabla de parmetros 2.6.7.Table 1-8 en el manual de instrucciones.)

Unidades Unidades de medida del sensor.

Opciones

M, CM, MM, FT, IN Predeterminado: M

Espaol

Nota: En este ejemplo se supone que todos los valores aparecen en metros (m).

7ML19985XU81

SITRANS LUT400 (HART)

Pgina ES -21

Punto de calibracin superior Distancia del punto de referencia del sensor al punto de calibracin superior: normalmente el nivel del depsito lleno..

Valor

Rango: 0,000 a 60.000 Predeterminado: 0,000

Punto de calibracin inferior Distancia del punto de referencia del sensor al punto de calibracin inferior: normalmente el nivel del depsito vaco..

Valor

Rango: 0,000 a 60.000 Predeterminado: 60.000

Velocidad de respuesta Determina la velocidad de reaccin del dispositivo a los cambios de medida en el rango objetivo.

Notas:

La velocidad de respuesta slo se puede determinar con el asistente, y cualquier cambio realizado en los parmetros Fill Rate per Minute (2.3.1.) o Empty Rate per Minute (2.3.2.) despus de haber terminado ste sobrescribir el valor de la velocidad de respuesta. Velocidad de respuesta siempre aparece en m/minuto.
SLOW (lento) (0,1 M/MIN) MEDIUM (medio) (1,0 M/MIN) FAST (rpido) (10 M/MIN) Predeterminado: SLOW (lento) (0,1 M/MIN)

Opciones

Use un valor ligeramente superior a la velocidad mxima de llenado o vaciado (cualquiera que sea mayor). Con una velocidad baja se consigue mayor precisin, y con una alta se registran ms fluctuaciones de nivel.

Dimensin A Altura de la base del depsito cuando es cnica, piramidal, parablica, esfrica o con inclinacin plana.

Valor

Rango: 0,000 a 99.999 Predeterminado: 0,000

Dimensin L Longitud de la seccin cilndrica de un depsito horizontal de extremo parablico.

Valor

Rango: 0,000 a 99.999 Predeterminado: 0,000

Unidades de volumen Determina las unidades de medida del volumen.

Espaol

Opciones

L, USGAL, IMPGAL, CUM, USER DEFINED (definido por usuario) * Predeterminado: L

* Si selecciona la opcin USER DEFINED (Definido por usuario), deber poner el valor cuando termine el asistente. Consulte los parmetros en la seccin 2.6.6.User Defined Unit del manual de instrucciones.

Pgina ES-22

SITRANS LUT400 (HART)

7ML19985XU81

Volumen mximo
Es el volumen mximo del depsito. Ponga el volumen del punto superior de calibracin. Por ejemplo: si el volumen mximo del depsito es 8.000l, ponga 8.000.

Valor

Rango: 0,0 a 9999999 Predeterminado: 100.0

Fin del asistente de QS Volume (volumen QS)


Para que QS funcione correctamente, hay que guardar todos los cambios.
BACK (anterior), CANCEL (cancelar), FINISH (terminar) [La pantalla vuelve al men 1.1 Quick Start (asistente) cuando el asistente ha terminado o se ha cancelado. Si se ha seleccionado CANCEL, no quedar registrado ningn cambio].

Opciones

Para transferir los valores del asistente al dispositivo y volver al men Program (programacin), pulse la flecha hacia ABAJO veces la flecha a la IZQUIERDA (Terminar). Pulse a continuacin tres para volver al modo de medicin.

1.1.3. Caudal QS
Use este asistente para configurar aplicaciones sencillas de caudal. [Visible slo en los modelos configurados de LUT430 (bombeo y caudal) y LUT440 (OCM)]. a) Pulse la flecha a la DERECHA para activar el modo PROGRAM (programacin) y abrir el nivel de mens 1: MEN PRINCIPAL b) Pulse dos veces la flecha a la DERECHA c) Pulse dos veces la flecha ABAJO (caudal QS). para ir al elemento 1.1.1. del men. para abrir el men QS Flow

y a la DERECHA

d) Pulse siempre la flecha hacia ABAJO para aceptar los valores predeterminados y pasar directamente al elemento siguiente, o la flecha a la DERECHA para abrir el modo de edicin: en pantalla aparece resaltada la seleccin. e) Desplcese hasta el elemento que le interesa, pulse la flecha a la DERECHA el cambio y pulse la flecha ABAJO continuar. f) para guardar para

ASSISTENTE PUESTA EN MARCHA RAP..


TRANSDUCTOR XRS-5
RETORNO ANULAR SIGUIENTE EDITAR

ASISTENTE PUESTA EN MARCHA RAPI..


TRANSDUCTOR

SIN TXT XPS-10 XPS-15 XCT-8 XCT-12

XRS-5

En cualquier momento puede pulsar la flecha hacia ARRIBA IZQUIERDA para cancelar el asistente.

para volver o a la

Inicio del asistente de QS Flow (caudal QS)


Muestra el tipo de asistente que se va a ejecutar.

Opciones

CANCEL (cancelar), START (inicio)

Espaol

7ML19985XU81

SITRANS LUT400 (HART)

Pgina ES -23

Transductor Determina qu transductor Siemens est conectado al dispositivo.

Opciones

NO TRANSDUCER (sin transductor), XRS-5, XPS-10, XPS-15, XCT-8, XCT-12, XPS-30, XPS-40, XLT-30, XLT-60, STH Predeterminado: NO TRANSDUCER

Fuente de temperatura Es la fuente de la lectura de temperatura que se va a usar para poner la velocidad del sonido.

Opciones

TRANSDUCER (transductor), FIXED TEMPERATURE (temperatura fija), EXTERNAL TS-3 (TS-3 externo), AVERAGE OF SENSORS (media de sensores) Predeterminado: TRANSDUCER

Temperatura fija Es la fuente de la lectura de temperatura que se va a usar para poner la velocidad del sonido.

Valor

Rango: -100,0 a +150,0 C Predeterminado: +20,0 C

Este parmetro slo aparece si est seleccionada la opcin FIXED TEMPERATURE (Temperatura fija) en Fuente de temperatura.

Dispositivo principal de medicin


Define el dispositivo principal de medicin (PMD) que se va a usar en la aplicacin.
EXPONENTIAL DEVICES (dispositivos exponenciales), RECTANGULAR FLUME BS-3680 (canal rectangular BS-3680), ROUND NOSE HORIZONTAL CR. BS-3680 (vertedero de cresta redondeada horizontal BS-3680, TRAPEZOIDAL FLUME BS-3680 (canal trapezoidal BS-3680), U-FLUME BS3680 (canal de corte en U BS-3680), FINITE CREST WEIR BS-3680 (vertedero de cresta limitada BS-3680), THIN PLATE RECT. WEIR BS-3680 (vertedero rectangular de placa fina BS-3680), THIN PLATE V-NOTCH WEIR BS-3680 (vertedero de placa fina con corte en V BS-3680) , RECT. WEIR CONTRACTED (vertedero rectangular con contraccin), ROUND PIPE (conducto redondo), PALMER BOWLUS FLUME (canal Palmer Bowlus), HFLUME (canal en H), OTHER (otros)* Predeterminado: EXPONENTIAL DEVICES (dispositivos exponenciales)

Opciones

* La opcin ser OTHER (otros) si el asistente se ejecut anteriormente en una aplicacin HART (como SIMATIC PDM), y el dispositivo se puso en OFF o en UNIVERSAL HEAD VS. FLOW (Altura mxima universal frente a caudal). Si es sta la configuracin inicial, el dispositivo principal slo se puede poner para que no calcule (en OFF) o para linealizacin [UNIVERSAL HEAD VS. FLOW (altura mxima universal frente a caudal)] mediante la aplicacin HART (SIMATIC PDM, AMS, FC375/475). Consulte el manual de instrucciones.

Mtodo para calcular el caudal Determina con qu mtodo se va a calcular el caudal.

Espaol

Opciones

ABSOLUTE (absoluto), RATIOMETRIC (ratiomtrico) Predeterminado: ABSOLUTE

Pgina ES-24

SITRANS LUT400 (HART)

7ML19985XU81

Unidades Unidades de medida del sensor.

Opciones

M, CM, MM, FT, IN Predeterminado: M

Nota: En este ejemplo se supone que todos los valores aparecen en metros (m). Punto de calibracin superior Distancia del punto de referencia del sensor al punto de calibracin superior: normalmente el nivel del depsito lleno..

Valor

Rango: 0,000 a 60.000 Predeterminado: 0.000

Punto de calibracin inferior Distancia del punto de referencia del sensor al punto de calibracin inferior: normalmente el nivel del depsito vaco..

Valor

Rango: 0,000 a 60.000 Predeterminado: 60.000

Velocidad de respuesta Determina la velocidad de reaccin del dispositivo a los cambios de medida en el rango objetivo. Notas:
La velocidad de respuesta slo se puede determinar con el asistente, y cualquier cambio realizado en los parmetros Fill Rate per Minute (2.3.1.) o Empty Rate per Minute (2.3.2.) despus de haber terminado ste sobrescribir el valor de la velocidad de respuesta. Velocidad de respuesta siempre aparece en m/minuto.
SLOW (lento) (0,1 M/MIN) MEDIUM (medio) (1,0 M/MIN) FAST (rpido) (10 M/MIN) Predeterminado: SLOW (lento) (0,1 M/MIN)

Opciones

Use un valor ligeramente superior a la velocidad mxima de llenado o vaciado (cualquiera que sea mayor). Con una velocidad baja se consigue mayor precisin, y con una alta se registran ms fluctuaciones de nivel.

Factores de clculo:

Notas: En el asistente aparecern los cinco parmetros siguientes segn lo que se haya seleccionado como dispositivo principal. Estos parmetros se usan en las frmulas de clculo del caudal (para ms informacin, consulte el manual de instrucciones).

Exponente de caudal
[PMD = EXPONENTIAL DEVICES (Dispositivos exponenciales)]

Espaol

Es el exponente que se usa en la frmula de clculo del caudal.

Valor
7ML19985XU81

Rango: -999.000 a 9999.000 Predeterminado: 1.550

SITRANS LUT400 (HART)

Pgina ES -25

Factor K
[PMD = EXPONENTIAL DEVICES (Dispositivos exponenciales)]

Es la constante que se usa en las frmulas de clculo de caudal, slo en un dispositivo exponencial.

Valor
Ranura en V

Rango: -999.000 a 9999.000 Predeterminado: 1.000

[PMD = THIN PLATE V-NOTCH WEIR (Vertedero de placa fina con corte en V)]

El corte en V se usa en la frmula de clculo del caudal.

Valor
Pendiente

Rango: 25.000 a 95.000 Predeterminado: 25.000

[PMD = TRAPEZOIDAL FLUME (Canal trapezoidal) o ROUND PIPE (Conducto redondo)]

Es la pendiente que se usa en la frmula de clculo de caudal.

Valor

Rango: -999.000 a 9999.000 Predeterminado: 0.000

Coeficiente de rugosidad
[PMD = ROUND PIPE (Conducto redondo)]

El coeficiente de rugosidad se usa en la frmula de clculo del caudal.

Valor

Rango: -999.000 a 9999.000 Predeterminado: 0.000

Espaol
Pgina ES-26

SITRANS LUT400 (HART)

7ML19985XU81

Dimensiones del dispositivo principal


Notas: Para cada dispositivo principal, salvo los Exponential Devices (Dispositivos exponenciales) y Other (Otros), tiene que poner hasta cuatro dimensiones. El asistente le pedir todas las dimensiones del dispositivo principal seleccionado, presentando a continuacin el nombre de la dimensin correspondiente. PMD seleccionado Nombre de la dimensin en el asistente (men de parmetros)
APPROACH WIDTH (ancho del canal de acceso) B (2.15.4.5. OCM Dimension 1) THROAT WIDTH (ancho de la garganta) B (2.15.4.6. OCM Dimension 2) HUMP HEIGHT (altura del lomo) P (2.15.4.7. OCM Dimension 3) THROAT LENGTH (ancho de la garganta) L (2.15.4.8. OCM Dimension 4)

Rectangular Flume BS-3680 (canal rectangular BS-3680)

Round Nose Horizontal Crest Weir BS-3680 (vertedero de cresta redondeada horizontal BS-3680)
CREST WIDTH (ancho de la cresta) B(2.15.4.5. OCM Dimension 1) CREST HEIGHT (alto de la cresta) P (2.15.4.6. OCM Dimension 2) CREST LENGTH (longitud de la cresta) L (2.15.4.7. OCM Dimension 3)

Trapezoidal Flume BS-3680 (canal trapezoidal BS-3680)


Dimension 1)
APPROACH WIDTH (ancho del canal de acceso) B (2.15.4.5. OCM THROAT WIDTH (ancho de la garganta) B (2.15.4.6. OCM Dimension 2) HUMP HEIGHT (altura del lomo) P (2.15.4.7. OCM Dimension 3) THROAT LENGTH (ancho de la garganta) L (2.15.4.8. OCM Dimension 4)

U-Flume BS-3680 (canal con corte en U BS-3680)


Dimension 1)
APPROACH DIAMETER (dimetro de alcance) DA (2.15.4.5. OCM THROAT DIAMETER (dimetro de la garganta) D (2.15.4.6. OCM Dimension 2) HUMP HEIGHT (altura del lomo) P (2.15.4.7. OCM Dimension 3) THROAT LENGTH (ancho de la garganta) L (2.15.4.8. OCM Dimension 4)

Finite Crest Weir BS-3680 (vertedero de cresta limitada BS-3680)


CREST WIDTH (ancho de la cresta) B(2.15.4.5. OCM Dimension 1) CREST HEIGHT (alto de la cresta) P (2.15.4.6. OCM Dimension 2) CREST LENGTH (longitud de la cresta) L (2.15.4.7. OCM Dimension 3)

Thin Plate Rectangular Weir BS-3680 (vertedero rectangular de placa fina BS-3680)
Dimension 1)
APPROACH WIDTH (ancho del canal de acceso) B (2.15.4.5. OCM CREST WIDTH (ancho de la cresta) B(2.15.4.6. OCM Dimension 2)

Espaol

CREST HEIGHT (alto de la cresta) P (2.15.4.7. OCM Dimension 3)

Rectangular Weir Contracted (vertedero rectangular con contraccin)


CREST WIDTH (ancho de la cresta) B (2.15.4.5. OCM Dimension 1) (contina en la pgina siguiente)

7ML19985XU81

SITRANS LUT400 (HART)

Pgina ES -27

PMD seleccionado (continuacin)

Nombre de la dimensin en el asistente (men de parmetros)


PIPE INSIDE DIAMETER (dimetro interior del conducto) D (2.15.4.5. OCM Dimension 1)

Round Pipe (conducto redondo)

Palmer Bowlus Flume (canal Palmer Bowlus)


MAXIMUM FLUME WIDTH HMAX (ancho mximo del canal, altura mxima) ((2.15.4.5. OCM Dimension 1)

H-Flume (canal con corte en H)


MAXIMUM LISTED HEAD HMAX (altura mxima registrada, altura mxima) (2.15.4.5. OCM Dimension 1)

Altura mxima Es el nivel mximo del dispositivo principal.

Valor

Rango: 0,000 a 60.000 Predeterminado: 60.000

Desviacin de la altura mxima cero Es la diferencia (positiva o negativa) entre Punto de calibracin inferior y la altura mxima cero (nivel con un caudal cero).

Valor

Rango: -60.000 a 60.000 Predeterminado: 0.000

Unidades de caudal Son las unidades de volumen que se usan para presentar el flujo total.

Opciones

L/S, L/MIN, CUFT/S, CUFT/D, GAL/MIN, GAL/D, IMPGAL/MIN, IMPGAL/D, CUM/H, CUM/D, USER DEFINED (definido por usuario) * Predeterminado: L/S

* Si selecciona la opcin USER DEFINED (Definido por usuario), deber poner el valor cuando termine el asistente. Consulte los parmetros en la seccin 2.6.6.User Defined Unit del manual de instrucciones.

Caudal mximo a 20 mA Es el caudal mximo.

Valor

Rango: -999 a 9999999 Predeterminado: 100

Decimales del caudal Es el nmero mximo de decimales que van a aparecer.

Opciones

NO DIGITS (ninguno), 1 DIGIT (1 decimal), 2 DIGITS (2 decimales), 3 DIGITS (3 decimales) Predeterminado: NO DIGITS

Espaol

Interrupcin por caudal bajo Elimina la actividad del totalizador cuando el nivel mximo llega al valor de interrupcin o queda por debajo de l.

Valor

Rango: 0,000 a 60.000 Predeterminado: 0.000

Pgina ES-28

SITRANS LUT400 (HART)

7ML19985XU81

Fin del asistente de QS Flow (caudal QS)


Para que QS funcione correctamente, hay que guardar todos los cambios.
BACK (anterior), CANCEL (cancelar), FINISH (terminar) [La pantalla vuelve al men 1.1.1 Quick Start (asistente) cuando el asistente ha terminado o se ha cancelado. Si se ha seleccionado CANCEL, no quedar registrado ningn cambio].

Opciones

Para transferir los valores del asistente al dispositivo y volver al men Program (programacin), pulse la flecha hacia ABAJO veces la flecha a la IZQUIERDA (Terminar). Pulse a continuacin tres para volver al modo de medicin.

Nota: Es altamente recomendable realizar una 2.15.2.Auto Zero Head despus de terminar el asistente para asegurar la mxima precisin. Consulte los detalles en el manual de instrucciones.

1.2. Control de bombas


Si la aplicacin incluye el uso de bombas, utilice este asistente para configurarlas. Termine de ejecutar primero el asistente correspondiente. a) Pulse la flecha a la DERECHA para activar el modo PROGRAM (Programacin) y abrir el nivel de mens 1: MEN PRINCIPAL b) Pulse la flecha a la DERECHA c) Pulse la flecha ABAJO de bombas). d) Pulse siempre la flecha hacia ABAJO para aceptar los valores predeterminados y pasar directamente al elemento siguiente, o pulse la flecha a la DERECHA resaltada la seleccin. para abrir el modo de edicin: en pantalla aparece para para ir al elemento 1.1. del men. para abrir el men Pump Control (Control

y a la DERECHA

e) Desplcese hasta el elemento que le interesa, pulse la flecha a la DERECHA guardar el cambio y ABAJO f) para continuar.

En cualquier momento puede pulsar la flecha hacia ARRIBA IZQUIERDA para cancelar el asistente.

para volver o a la

Inicio del asistente Pump Control (control de bombas)


Muestra el tipo de asistente que se va a ejecutar.

Opciones

CANCEL (cancelar), START (inicio)

Nmero de bombas Seleccione el nmero de bombas que se van a usar con control.

Opciones

NONE (ninguna), 2 Predeterminado: NONE

Espaol

Si est puesto en NONE (Ninguna), el control est desactivado.

7ML19985XU81

SITRANS LUT400 (HART)

Pgina ES -29

Relay Pump 1 (rel de bomba 1) Selecciona el rel asignado a la bomba 1.

Opciones

RELAY 2 (rel 2), RELAY 3 (rel 3) Predeterminado: RELAY 2

Relay Pump 2 (rel de bomba 2) Slo visualizacin. Determina automticamente el rel asignado a la bomba 2 basndose en un rel asignado a la bomba 1 en un paso anterior. Opciones 2 (rel de bomba 2) = RELAY 3 (rel 3) (slo Si Relay Pump 1 (rel de bomba 1) = RELAY 3 (rel 3), entonces Relay Pump visualizacin)
2 (rel de bomba 2) = RELAY 2 (rel 2) Si Relay Pump 1 (rel de bomba 1) = RELAY 2 (rel 2), entonces Relay Pump

Modo Pump Control (control de bombas)


Determina el algoritmo de control usado para activar el rel.

Controlador de nivel LUT420:


ALTERNATE DUTY ASSIST (Ayuda alternativa de funcionamiento), ALTERNATE DUTY BACKUP (Copia de seguridad alternativa)

Controlador de bombeo y caudal LUT430:


ALTERNATE DUTY ASSIST (Ayuda alternativa de funcionamiento), ALTERNATE DUTY BACKUP (Copia de seguridad alternativa), SERVICE RATIO DUTY ASSIST (Ayuda de funcionamiento velocidad de servicio), SERVICE RATIO DUTY BACKUP (Copia de seguridad velocidad de servicio), FIXED DUTY ASSIST (Ayuda fija de funcionamiento), FIXED DUTY BACKUP (Copia de seguridad fija)

Opciones compatibles con cada modelo

OCM de alta precisin LUT440:


ALTERNATE DUTY ASSIST (Ayuda alternativa de funcionamiento), ALTERNATE DUTY BACKUP (Copia de seguridad alternativa), SERVICE RATIO DUTY ASSIST (Ayuda de funcionamiento velocidad de servicio), SERVICE RATIO DUTY BACKUP (Copia de seguridad velocidad de servicio), FIXED DUTY ASSIST (Ayuda fija de funcionamiento), FIXED DUTY BACKUP (Copia de seguridad fija) Predeterminado (todos los modelos): ALTERNATE DUTY ASSIST

Velocidad de funcionamiento de la bomba 1 Selecciona el uso de la bomba basndose en la velocidad de funcionamiento y no en la ltima velocidad utilizada.

Valor

Rango: 0 a 255 Predeterminado: 1

Este parmetro slo aparece si se selecciona un algoritmo en el modo Pump Control (Control de bombas).

Velocidad de funcionamiento de la bomba 2 Selecciona el uso de la bomba basndose en la velocidad de funcionamiento y no en la ltima velocidad utilizada..

Espaol

Valor

Rango: 0 a 255 Predeterminado: 1

Este parmetro slo aparece si se selecciona un algoritmo en el modo Pump Control (Control de bombas).

Pgina ES-30

SITRANS LUT400 (HART)

7ML19985XU81

Tiempo de funcionamiento del rel 2 Defina en horas el tiempo que ha transcurrido para el rel de la bomba 2.

Valor

Rango: 0 a 999999 Predeterminado: 0

Ponga el valor predeterminado si las bombas son nuevas o para bombas ya existentes con tiempo de funcionamiento acumulado. [Este parmetro slo aparece si se selecciona un algoritmo en el modo Pump Control (Control de bombas)].

Tiempo de funcionamiento del rel 3 Defina en horas el tiempo que ha transcurrido para el rel de la bomba 3.

Valor

Rango: 0 a 999999 Predeterminado: 0

Ponga el valor predeterminado si las bombas son nuevas o para bombas ya existentes con tiempo de funcionamiento acumulado. [Este parmetro slo aparece si se selecciona un algoritmo en el modo Pump Control (Control de bombas)].

Nivel de conexin bomba 1 Es el nivel en el que se conecta la bomba 1, definido en 2.1.1.Units.

Valor

Rango: 0,000 a 99999.000 Predeterminado: 0.000

Nivel de conexin bomba 2 Es el nivel en el que se conecta la bomba 2, definido en 2.1.1.Units.

Valor

Rango: 0,000 a 99999.000 Predeterminado: 0.000

Nivel de desconexin bomba 1 Es el nivel en el que se desconecta la bomba 1, definido en 2.1.1.Units.

Valor

Rango: 0,000 a 99999.000 Predeterminado: 0.000

Nivel de desconexin bomba 2 Es el nivel en el que se desconecta la bomba 2, definido en 2.1.1.Units.

Valor

Rango: 0,000 a 99999.000 Predeterminado: 0.000

Fin del asistente - Control de bombas


Para que el asistente funcione correctamente, hay que guardar todos los cambios.
BACK (anterior), CANCEL (cancelar), FINISH (terminar) [La pantalla vuelve al men Pump Control (control de bombas) cuando el asistente ha terminado o se ha cancelado). Si se ha seleccionado CANCEL, no quedar registrado ningn cambio].

Opciones

Para transferir los valores del asistente al dispositivo y volver al men Program (programacin), pulse la flecha hacia ABAJO veces la flecha a la IZQUIERDA (Terminar). Pulse a continuacin dos para volver al modo de medicin.

Espaol

7ML19985XU81

SITRANS LUT400 (HART)

Pgina ES -31

Comunicaciones con SITRANS LUT400: HART


SITRANS LUT400 se puede configurar en la red HART mediante la aplicacin HART Communicator 375/475 de Emerson (para ms informacin, consulte el manual de instrucciones), o mediante un paquete de software. El paquete que recomendamos es SIMATIC Process Device Manager (PDM) de Siemens. En la gua Working with HART (Manual de HART), que se puede descargar en la pgina de productos de nuestro sitio web, podr ver ms detalles. Vaya a: www.siemens.com/sitransLUT400, seccin Support (Servicio tcnico) y haga clic en Application Guides (Manuales de aplicaciones).

Funcionamiento remoto
SITRANS LUT400 es compatible con varias herramientas informticas de comunicacin remota: PC con SIMATIC PDM instalado PC con Emerson AMS Device Manager instalado PC con un navegador de Internet instalado PC con una aplicacin para dispositivos de campo (FDT) instalada Field Communicator 375/475 (FC375/FC475). Ms abajo encontrar la informacin bsica para usar estas aplicaciones con SITRANS LUT400. En el anexo sobre comunicaciones del manual SITRANS LUT400 (HART) encontrar ms informacin.

SIMATIC PDM
SIMATIC PDM es un paquete de software que se usa para poner en marcha y mantener SITRANS LUT400 y otros dispositivos de procesos. En www.siemens.com/simatic-pdm podr ver ms informacin. Consulte la pgina de asistencia de nuestro sitio web para ver si tiene la ltima versin de SIMATIC PDM, el Service Pack (SP) y el hot fix (HF) (parche) ms recientes. Vaya a: http://support.automation.siemens.com/WW/ llisapi.dll?func=cslib.csinfo&lang=en&objiD=10806857&subtype=133100

Update (actualizar) Electronic Device Description (descripcin de dispositivos electrnicos) (EDD)


Puede encontrar la EDD en el Device Catalog (catlogo de dispositivos), en Sensors (Sensores)/ Level (Nivel)/Echo (Eco)/Siemens AG/SITRANS LUT400. (La EDD est diseada para ser compatible con versiones posteriores). Como gua para encontrar la EDD correcta, los nmeros mayor y menor debern estar entre la revisin de EDD y la revisin de Firmware del dispositivo (por ejemplo, el nmero mayor y menor en negrita: 1.00.00-02).
La revisin de EDD tiene que coincidir con la revisin de Firmware del dispositivo. Para verlo en PDM, vaya a SITRANS LUT400 HART > Identification (Identificacin) > Device (Dispositivo).

misma revisin de Firmware y de EDD

La instalacin de una nueva versin de SIMATIC PDM requiere el ltimo Service Pack (SP) y el hot fix (parche) (HF) ms reciente.

Cmo instalar una nueva EDD


Vaya a www.siemens.com/sitransLUT400 > Support (Servicio tcnico) > Software Downloads (Descargas de software) para descargar la EDD ms actual de la pgina de productos de nuestro sitio web. Guarde los archivos y descomprmalos en una carpeta de fcil acceso. Abra el SIMATIC PDM - Manage Device Catalog (Catlogo de dispositivos de gestin) y seleccione la carpeta que contiene el archivo comprimido EDD.

Espaol

Pgina ES-32

SITRANS LUT400 (HART)

7ML19985XU81

Cmo configurar un nuevo dispositivo


1) 2) Mire primero si tiene la ltima EDD; de no ser as, actualcela (consulte ms arriba la seccin Cmo instalar una nueva EDD). Configure SITRANS LUT400 con el asistente que corresponda a su aplicacin.

Asistentes en SIMATIC PDM


Notas: Los parmetros del asistente estn interrelacionados, y los cambios slo se harn efectivos si hace clic en Apply and Transfer (Aplicar y transferir) despus de haber dado el ltimo paso. Haga clic en BACK (Anterior) para volver y revisar un parmetro o en CANCEL (Cancelar) para salir del asistente. Abra el men Device (Dispositivo) de SIMATIC PDM y seleccione el Quick Start (Asistente) de su aplicacin [Level (Nivel), Volume (Volumen), Volume-Linearization (Linealizacin del volumen) o Flow (Caudal)].

Asistentes

Funcionamiento con la aplicacin AMS Device Manager


AMS Device Manager es un paquete de software diseado para poner en marcha y mantener SITRANS LUT400 y otros dispositivos de procesos. Si desea ms informacin, vaya a: http://www.emersonprocess.com/AMS/.

Electronic Device Description (descripcin de dispositivos electrnicos, EDD)


SITRANS LUT400 requiere la EDD para AMS Device Manager, versin 10.5.

Configuracin de un nuevo dispositivo con AMS Device Manager


1) Puede encontrar la EDD en el Device Catalog (catlogo de dispositivos), en Sensors (Sensores)/ Level (Nivel)/Echo (Eco)/Siemens AG/SITRANS LUT400. Mire la pgina de productos de nuestro sitio web en www.siemens.com/sitransLUT400, seccin Support (Servicio tcnico) >Software Downloads (Descargas de software), para ver si tiene la ltima versin de la EDD para AMS Device Manager. Guarde los archivos y descomprmalos en una carpeta de fcil acceso. Abra AMS Device Manager - Add Device Type (Agregar tipo de dispositivo), busque el archivo EDD comprimido y seleccinelo. [En la pgina de productos, seccin Support (Servicio tcnico) de nuestro sitio web podr descargar una Application Guide (Gua de aplicaciones) para configurar dispositivos HART con AMS Device Manager]. En Device Connection View (Vista de conexiones de dispositivos), haga clic con el botn derecho del ratn sobre el icono del dispositivo y seleccione Scan Device (Ver dispositivo) para ver los parmetros del dispositivo. Haga doble clic en el icono del dispositivo para abrir la pantalla de inicio. Esta pantalla muestra los datos de identificacin del dispositivo y una ventana de navegacin en la parte izquierda de la pantalla. Configure SITRANS LUT400 con el asistente correspondiente.

2) 3)

4)

5)

Espaol

6)

7ML19985XU81

SITRANS LUT400 (HART)

Pgina ES -33

Funcionamiento con FC375/475 (HART)


FC375/FC475 HART Communicator es un dispositivo porttil de comunicacin fcil de usar y compatible con cualquier dispositivo HART como SITRANS LUT400. La estructura de mens de FC375/ 475 es muy similar a la de AMS Device Manager.

Electronic Device Description (descripcin de dispositivos electrnicos, EDD)


SITRANS LUT400 requiere la EDD de FC375/FC475.

Configuracin de un nuevo dispositivo con FC375/FC475


1) Mire la pgina de productos de nuestro sitio web www.siemens.com/sitransLUT400, en la seccin Support (Servicio tcnico)>Software Downloads (Descargas de software), para ver si tiene la ltima versin de la EDD de FC375/FC475. Guarde los archivos y descomprmalos en una carpeta de fcil acceso. Para instalar la EDD, ejecute la Field Communicator Easy Upgrade Utility (herramienta de actualizacin simplificada de comunicador de campo) de Emerson. Configure SITRANS LUT400 con el asistente que corresponda a su aplicacin.

2) 3) 4)

Funcionamiento con FDT [Field Device Tool (Herramienta de dispositivo de campo)]


FDT es un estndar que se usa en diversos paquetes de software diseados para poner en marcha y mantener dispositivos de campo como SITRANS LUT400. Dos FDT de venta al pblico son PACTware y Fieldcare. FDT es muy similar a PDM. Para configurar un dispositivo de campo con FDT se necesita su correspondiente DTM [Device Type Manager (Gestor de tipos de dispositivos)]. Siemens Instruments usa SITRANS DTM y una EDD creada para SITRANS DTM. 1) Empiece por instalar SITRANS DTM. Si lo desea, lo puede descargar en: http://www.siemens.com/sitransdtm. Haga clic en Support (Servicio tcnico) y vaya a Software Downloads (Descargas de software). Instale la EDD de SITRANS LUT400 (HART) para SITRANS DTM. La puede descargar en la pgina de productos de nuestro sitio web: Vaya a www.siemens.com/sitransLUT400, seccin Support (Servicio tcnico) y haga clic en Software Downloads (Descargas de software).

2)

Configuracin de un nuevo dispositivo con FDT


En la seccin Support (Servicio tcnico) de la pgina de productos de nuestro sitio web podr descargar una gua de aplicaciones.

Servicio y mantenimiento
Si funciona en condiciones normales, SITRANS LUT400 no necesita mantenimiento ni limpieza.

Reparacin del dispositivo y lmite de responsabilidad


Si desea ms informacin, consulte la contracubierta de este documento.

Espaol
Pgina ES-34

SITRANS LUT400 (HART)

7ML19985XU81

Configuracin de conexiones en instalaciones de reas peligrosas


Para instalaciones en reas peligrosas existen las siguientes opciones de conexin: Conexiones a prueba de ignicin de polvo (Internacional/Europa) Conexiones no inflamables (Canad) Mire la etiqueta de su dispositivo para comprobar las homologaciones.

1. Conexiones a prueba de ignicin de polvo (Internacional/ Europa)


SITRANS LUT420 7MLxxxx-xxxxx-xxxx Serial No.: GYZ / B1034567 0518 KCC-REM-S49 Power Rating: 100 230V 15% SITRANSLUT400 50/60 Hz, 36 VA (10 W) Contact Rating: 1A/5A @ 250V , Non-Inductive Operating Temperature: 20C to 50C Enclosure: IP65 / TYPE 4X / NEMA 4X II 3D Ex tc IIIC T70C Dc Ex tc IIIC T70C Dc SIRA 12ATEX9143X IECEx SIR 12.0057X Warning: Possible static hazard. Do not rub or clean on site with a dry cloth. Warning: Do not open when energized or when an explosive dust atmosphere is present. Siemens Milltronics Process Instruments, Peterborough Assembled in Canada with domestic and imported parts SITRANS LUT430 7MLxxxx-xxxxx-xxxx Serial No.: GYZ / B1034567 0518 KCC-REM-S49 Power Rating: 100 230V 15% SITRANSLUT400 50/60 Hz, 36 VA (10 W) Contact Rating: 1A/5A @ 250V , Non-Inductive Operating Temperature: 20C to 50C Enclosure: IP65 / TYPE 4X / NEMA 4X II 3D Ex tc IIIC T70C Dc Ex tc IIIC T70C Dc SIRA 12ATEX9143X IECEx SIR 12.0057X Warning: Possible static hazard. Do not rub or clean on site with a dry cloth. Warning: Do not open when energized or when an explosive dust atmosphere is present. Siemens Milltronics Process Instruments, Peterborough Assembled in Canada with domestic and imported parts SITRANS LUT440 7MLxxxx-xxxxx-xxxx Serial No.: GYZ / B1034567 0518 KCC-REM-S49 Power Rating: 100 230V 15% SITRANSLUT400 50/60 Hz, 36 VA (10 W) Contact Rating: 1A/5A @ 250V , Non-Inductive Operating Temperature: 20C to 50C Enclosure: IP65 / TYPE 4X / NEMA 4X II 3D Ex tc IIIC T70C Dc Ex tc IIIC T70C Dc SIRA 12ATEX9143X IECEx SIR 12.0057X Warning: Possible static hazard. Do not rub or clean on site with a dry cloth. Warning: Do not open when energized or when an explosive dust atmosphere is present. Siemens Milltronics Process Instruments, Peterborough Assembled in Canada with domestic and imported parts

SITRANS LUT420 7MLxxxx-xxxxx-xxxx Serial No.: GYZ / B1034567 0518 KCC-REM-S49 Power Rating: 10 32V , 10W SITRANSLUT400 Contact Rating: 1A/5A @ 250V , Non-Inductive Operating Temperature: 20C to 50C Enclosure: IP65 / TYPE 4X / NEMA 4X

SITRANS LUT430 7MLxxxx-xxxxx-xxxx Serial No.: GYZ / B1034567 0518 KCC-REM-S49 Power Rating: 10 32V , 10W SITRANSLUT400 Contact Rating: 1A/5A @ 250V , Non-Inductive Operating Temperature: 20C to 50C Enclosure: IP65 / TYPE 4X / NEMA 4X

SITRANS LUT440 7MLxxxx-xxxxx-xxxx Serial No.: GYZ / B1034567 0518 KCC-REM-S49 Power Rating: 10 32V , 10W SITRANSLUT400 Contact Rating: 1A/5A @ 250V , Non-Inductive Operating Temperature: 20C to 50C Enclosure: IP65 / TYPE 4X / NEMA 4X

II 3D Ex tc IIIC T70C Dc Ex tc IIIC T70C Dc SIRA 12ATEX9143X IECEx SIR 12.0057X Warning: Possible static hazard. Do not rub or clean on site with a dry cloth. Warning: Do not open when energized or when an explosive dust atmosphere is present.

II 3D Ex tc IIIC T70C Dc Ex tc IIIC T70C Dc SIRA 12ATEX9143X IECEx SIR 12.0057X Warning: Possible static hazard. Do not rub or clean on site with a dry cloth. Warning: Do not open when energized or when an explosive dust atmosphere is present.

II 3D Ex tc IIIC T70C Dc Ex tc IIIC T70C Dc SIRA 12ATEX9143X IECEx SIR 12.0057X Warning: Possible static hazard. Do not rub or clean on site with a dry cloth. Warning: Do not open when energized or when an explosive dust atmosphere is present.

Siemens Milltronics Process Instruments, Peterborough Assembled in Canada with domestic and imported parts

Siemens Milltronics Process Instruments, Peterborough Assembled in Canada with domestic and imported parts

Siemens Milltronics Process Instruments, Peterborough Assembled in Canada with domestic and imported parts

El certificado ATEX que aparece en la etiqueta se puede descargar en la pgina de productos de nuestro sitio web www.siemens.com/sitransLUT400. El certificado IECEx se puede ver en el sitio web de IECEx. Vaya a: http://iecex.iec.ch y haga clic en Ex Equipment Certificates of Conformity y ponga el nmero del certificado (IECEx SIR Observe las normativas locales al hacer las conexiones. Consulte asimismo el certificado ATEX anteriormente mencionado.

Espaol

7ML19985XU81

SITRANS LUT400 (HART)

Pgina ES -35

2. Conexiones no inflamables (Canad)


SITRANS LUT420 7MLxxxx-xxxxx-xxxx Serial No.: GYZ / B1034567 Power Rating: 100 230V 15% 50/60 Hz, 36 VA (10 W) Contact Rating: 1A/5A @ 250V , Non-Inductive Operating Temperature: 20C to 50C Enclosure: IP65 / TYPE 4X / NEMA 4X SITRANS LUT430 7MLxxxx-xxxxx-xxxx Serial No.: GYZ / B1034567 Power Rating: 100 230V 15% 50/60 Hz, 36 VA (10 W) Contact Rating: 1A/5A @ 250V , Non-Inductive Operating Temperature: 20C to 50C Enclosure: IP65 / TYPE 4X / NEMA 4X SITRANS LUT440 7MLxxxx-xxxxx-xxxx Serial No.: GYZ / B1034567 Power Rating: 100 230V 15% 50/60 Hz, 36 VA (10 W) Contact Rating: 1A/5A @ 250V , Non-Inductive Operating Temperature: 20C to 50C Enclosure: IP65 / TYPE 4X / NEMA 4X

159134

Class I, Div.2, Gr. A, B, C & D T3C Class II, Div. 2, Gr. F & G Class III Per Drawing A5E03936871

159134

Class I, Div.2, Gr. A, B, C & D T3C Class II, Div. 2, Gr. F & G Class III Per Drawing A5E03936871

159134

Class I, Div.2, Gr. A, B, C & D T3C Class II, Div. 2, Gr. F & G Class III Per Drawing A5E03936871

Siemens Milltronics Process Instruments, Peterborough Assembled in Canada with domestic and imported parts

Siemens Milltronics Process Instruments, Peterborough Assembled in Canada with domestic and imported parts

Siemens Milltronics Process Instruments, Peterborough Assembled in Canada with domestic and imported parts

SITRANS LUT420 7MLxxxx-xxxxx-xxxx Serial No.: GYZ / B1034567 Power Rating: 10 32V , 10W Contact Rating: 1A/5A @ 250V , Non-Inductive Operating Temperature: 20C to 50C Enclosure: IP65 / TYPE 4X / NEMA 4X

SITRANS LUT430 7MLxxxx-xxxxx-xxxx Serial No.: GYZ / B1034567 Power Rating: 10 32V , 10W Contact Rating: 1A/5A @ 250V , Non-Inductive Operating Temperature: 20C to 50C Enclosure: IP65 / TYPE 4X / NEMA 4X

SITRANS LUT440 7MLxxxx-xxxxx-xxxx Serial No.: GYZ / B1034567 Power Rating: 10 32V , 10W Contact Rating: 1A/5A @ 250V , Non-Inductive Operating Temperature: 20C to 50C Enclosure: IP65 / TYPE 4X / NEMA 4X

159134

Class I, Div.2, Gr. A, B, C & D T3C Class II, Div. 2, Gr. F & G Class III Per DWG. A5E03936871

159134

Class I, Div.2, Gr. A, B, C & D T3C Class II, Div. 2, Gr. F & G Class III Per DWG. A5E03936871

159134

Class I, Div.2, Gr. A, B, C & D T3C Class II, Div. 2, Gr. F & G Class III Per DWG. A5E03936871

Siemens Milltronics Process Instruments, Peterborough Assembled in Canada with domestic and imported parts

Siemens Milltronics Process Instruments, Peterborough Assembled in Canada with domestic and imported parts

Siemens Milltronics Process Instruments, Peterborough Assembled in Canada with domestic and imported parts

El diagrama de conexiones nmero A5E03936871 de la CSA clase I, div 2, se puede descargar en la pgina de productos de nuestro sitio web www.siemens.com/sitransLUT400.

Instrucciones especficas para instalaciones en reas peligrosas (Referencia: directiva europea ATEX 94/9/EC, anexo II, 1/0/6)
Estas instrucciones se refieren al dispositivo que ha obtenido el certificado nmero SIRA 12ATEXT9143X: 1) 2) 3) 4) Si desea ms informacin sobre la utilizacin y el montaje, consulte el manual de instrucciones. Este equipo est clasificado como dispositivo de la categora 3D. Puede usarse en reas peligrosas de tipo 22. Este equipo tiene una temperatura superficial mxima de T70 C (en un ambiente a +50 C). Consulte el cdigo de prcticas correspondiente para seleccionar el dispositivo adecuado a la temperatura de ignicin de polvo que vaya a tener la aplicacin. El dispositivo est certificado para trabajar a una temperatura ambiente entre -20 y +50 C. Todas las entradas de cables y tubos tienen que cumplir los requisitos de la directiva europea 94/9/EC para el grupo II, categora 3D y mantener la clasificacin IP global de la caja. El cable tiene que cumplir el cdigo de prcticas aplicable, adems de tener un aislamiento capaz de soportar la temperatura mxima superficial de la caja (T70 C). El dispositivo no se ha analizado como sistema de proteccin (como se indica en la Directiva 94/9/EC Anexo II, clusula 1.5). La instalacin e inspeccin de este aparato deber efectuarse por personal cualificado en conformidad con los cdigos de prcticas aplicables (EN 60079-14 y EN 60079-17 en Europa). La reparacin de este aparato deber realizarse por personal cualificado en conformidad con los cdigos de prcticas aplicables (EN 60079-19 en Europa). La integracin de componentes o la sustitucin de parte del aparato deber realizarse por personal cualificado, en conformidad con las indicaciones que aparecen en la documentacin del fabricante. La anulacin manual se puede realizar mediante el interruptor de desconexin situado en la instalacin del edificio.

5) 6) 7) 8) 9) 10)

Espaol

11)

12)

Pgina ES-36

SITRANS LUT400 (HART)

7ML19985XU81

CONDICIONES ESPECIALES PARA UN USO SEGURO


El sufijo 'X' en el nmero de certificado indica condiciones de seguridad especficas: Todas las entradas de cables y tubos tienen que cumplir los requisitos de la directiva europea 94/9/EC para el grupo II, categora 3D y mantener la clasificacin IP global de la caja. El cable tiene que cumplir el cdigo de prcticas aplicable, adems de tener un aislamiento capaz de soportar la temperatura mxima superficial de la caja (T70C). La caja es aislante y en ciertas condiciones extremas puede generar niveles de carga electrosttica suficientes para provocar una ignicin. El usuario ser responsable de no instalar el equipo donde se puedan dar ciertas condiciones (como vapor a alta presin) que puedan llegar a provocar la acumulacin de cargas electrostticas en superficies aislantes. Por otra parte, el equipo se deber limpiar siempre con un pao hmedo. El equipo se deber instalar en una zona de bajo riesgo de daos mecnicos o, de no ser posible, proveer una proteccin adicional adecuada para la aplicacin en la que se va a utilizar. Todos los controladores de la serie LUT400 (LUT420, LUT430, LUT440) usados con tapa montada remotamente sobre panel se instalarn en una caja con calificacin adecuada que cumpla los requisitos de la directiva europea 94/9/EC para el grupo II, categora 3D y calificacin IP65 o superior.

Espaol

7ML19985XU81

SITRANS LUT400 (HART)

Pgina ES -37

Notas

Espaol
Pgina ES-38

SITRANS LUT400 (HART)

7ML19985XU81

SITRANS LUT400 (HART) Manuel de mise en service rapide


Ce document dcrit les caractristiques et les fonctions essentielles du SITRANS LUT400 (HART 1)). Il est fortement recommand de se reporter la version intgrale des instructions de service pour accder lensemble des fonctions. La version intgrale des instructions de service SITRANS LUT400 (HART) ainsi que les instructions relatives la Communication sont disponibles sur la page produit, l'adresse : www.siemens.com/sitransLUT400. Pour obtenir une version imprime, contactez votre reprsentant Siemens. Toute question sur le contenu de ce document peut tre adresse : Siemens AG Siemens Milltronics Process Instruments 1954 Technology Drive, P .O. Box 4225 Peterborough, Ontario, Canada, K9J 7B1 veuillez nous crire : techpubs.smpi@siemens.com

Franais

Copyright Siemens AG 2012. Tous droits rservs


Nous encourageons les utilisateurs se procurer les exemplaires imprims de ces documents, ou les versions lectroniques prpares et valides par Siemens Milltronics Process Instruments. Siemens Milltronics Process Instruments ne pourra tre tenu responsable du contenu de toute reproduction totale ou partielle des versions imprimes ou lectroniques.

Clause de non-responsabilit
Le contenu de ce document a t vrifi pour garantir la conformit avec les caractristiques de l'instrument. Des divergences tant possibles, nous ne pouvons en aucun cas garantir la conformit totale. Ce document est rvis et actualis rgulirement pour inclure les nouvelles caractristiques. Nhsitez pas nous faire part de vos commentaires. Sous rserve de modifications techniques.

MILLTRONICS est une marque dpose de Siemens Milltronics Process Instruments.

Assistance technique

L'assistance technique est disponible 24h/24. Vous trouverez les coordonnes de l'interlocuteur Siemens Automatisation le plus proche l'adresse : www.siemens.com/automation/partner : Slectionner d'abord la langue d'affichage (franais). Cliquer sur l'onglet Partenaire pour slectionner la branche souhaite (par ex. : Technique d'Automatisation). Cliquer sur la liste droulante pour choisir un produit dans la gamme. Niveau : +Technique d'Automatisation > +Systmes sensoriels >+Instrumentation des procds > +Mesure de Niveau > +Procd continu. Slectionner Ultrason. Dbit : +Technique d'Automatisation > +Systmes sensoriels >+Instrumentation des procds > +Mesure du Dbit. Slectionner Ultrason. Slectionner un pays, puis une ville/rgion. Slectionner Technical Support sous Service. Vous trouverez l'assistance technique en ligne l'adresse : www.siemens.com/automation/ support-request Indiquer le nom de l'appareil (SITRANS LUT400) ou la rfrence de commande, et cliquer sur Rechercher, puis slectionner le type de produit. Cliquer sur Suite. Entrer un mot cl pour la requte. Accder la documentation souhaite en ligne ou cliquez sur Suivant pour crire aux spcialistes de l'Assistance Technique Siemens. Assistance Technique Siemens IA/DT: Tl.+49 (0)911 895 7222
1)

HARTest une marque dpose de HART Communication Foundation.

7ML19985XU81

SITRANS LUT400 (HART)

Page FR-1

Consignes de scurit
Franais
ll est important de respecter les consignes fournies dans ce document afin de garantir la scurit de l'utilisateur ou de tiers et la protection de l'instrument ou de tout quipement connect ce dernier. Chaque avertissement s'associe une explication dtaille du niveau de prcaution recommand pour chaque opration. Le symbole d'AVERTISSEMENT fait rfrence une mention sur le produit. Il signifie que la mort, des blessures corporelles graves et/ou des dommages matriels consquents peuvent se produire si les dispositions de scurit correspondantes ne sont pas respectes. Le symbole d'AVERTISSEMENT est utilis lorsqu'il n'y a pas de mention correspondante sur le produit. Il signifie que la mort, des blessures corporelles graves et/ou des dommages matriels consquents peuvent se produire si les dispositions de scurit correspondantes ne sont pas respectes. N.B. : information importante concernant linstrument ou une section particulire des instructions de service.

Conformit FCC
U.S.A. uniquement : Rgles FCC (Federal Communications Commission)
AVERTISSEMENT : tout(e) changement ou modification effectu(e) sans laccord pralable de Siemens Milltronics peut remettre en cause les droits dutilisation de l'appareil. Notes : Cet instrument est conforme aux limites imposes (alina 15 de la rglementation FCC, Classe A). Cette rglementation vise assurer une protection suffisante contre les interfrences nocives lorsque l'instrument est utilis dans un environnement commercial. Cet instrument gnre, utilise et peut mettre de lnergie frquence radio. Pour cette raison, il doit tre install et utilis suivant les instructions fournies, pour viter toute interfrence nocive aux communications radio. En cas d'utilisation en zone rsidentielle cet instrument peut provoquer des interfrences radio nocives. Il appartiendra lutilisateur de prendre en charge les consquences de ces dysfonctionnements.

Page FR-2

SITRANS LUT400 (HART)

7ML19985XU81

SITRANS LUT400
AVERTISSEMENT : pour garantir la scurit, le SITRANS LUT400 doit tre utilis suivant les instructions fournies dans ce document.

Franais

N.B. : cet instrument est conu pour une utilisation en milieu industriel. En cas d'utilisation en zone rsidentielle, ces instruments peuvent provoquer diffrentes interfrences radio. Les contrleurs de niveau ultrasons monopoint SITRANS LUT400 permettent la mesure en continu de liquides, de boues et de solides ainsi que le contrle de dbit en canal ouvert de trs haute prcision. Ces instruments compacts s'adaptent des plages de mesure tendues. Ils sont compatibles avec tous les transducteurs EchoMax de Siemens, sur des plages de mesure allant de 0,3 60 mtres (en fonction du transducteur). SITRANS LUT400 est associ une interface locale pour l'utilisateur (LUI), dote d'un afficheur rtro-clair, pour la programmation et l'accs aux paramtres par menu. Il dispose, entre autres, d'une interface web intuitive, et d'un large choix d'assistants ddis la mise en service, pour une utilisation Plug and Play. SITRANS LUT400 est compatible avec le protocole de communication HART et de nombreux logiciels pour la communication distance [SIMATIC PDM, AMS Device Manager, Field Communicator 375/475, et FDTs (Field Device Tools)]. Dot des techniques de traitement de signaux Sonic Intelligencede dernire gnration, il garantit un rendement sans gal et une grande facilit d'utilisation.

Caractristiques techniques
Alimentation
Version CA

Pour une liste exhaustive se reporter aux Instructions de service SITRANS LUT400 (HART).

100-230 V CA Fusible : 5 x 20 mm, action retarde 0,25A, 250V


Puissance absorbe maximale.

Version CC
10-32 V CC, 10W 1 Fusible : 5 x 20 mm, action retarde 1,6A, 125V

15%, 50 / 60 Hz, 36 VA (10W)1)

1)

Conditions environnementales
Altitude
max. 2000 m

Temprature ambiante
-20 +50 C (-4 +122 F)

Catgorie dinstallation
II

Degr de pollution
4

Humidit relative
Utilisable en extrieur (uniquement avec un botier IP65 Type 4X / NEMA 4X)

Homologations
Notes : Les homologations applicables sont indiques sur l'tiquette de l'appareil. Pour garantir la protection IP/NEMA veiller l'tanchit des conduits. Utilisation gnrale Zone dangereuse Non incendiaire (Canada) Atmosphre explosive (poussire) (International/Europe) CSAUS/C, CE, FM, UL listed, C-TICK CSA Classe I, Div. 2, Groupes A, B, C, D; Classe II, Div 2, Groupes F, G ; Classe III ATEX II 3D Ex tc IIIC T70 C IP65 IECEx SIR 12.0057X SIRA 12ATEX9143X Page FR-3

7ML19985XU81

SITRANS LUT400 (HART)

Installation
Franais
Notes : L'installation doit tre effectue par un personnel qualifi, en respectant les dispositions rglementaires locales. Cet instrument est sensible aux dcharges lectrostatiques. Assurer une mise la terre approprie. Tous les cblages doivent tre isols pour 250V minimum. Une tension dangereuse est prsente aux bornes de raccordement des transducteurs durant le fonctionnement : manipuler avec prcaution. Les bornes d'entre CC doivent tre alimentes par une source mme de fournir l'isolation lectrique entre/sortie requise pour la conformit avec les rgles de scurit de la norme CEI 61010-1. Les borniers de contact des relais doivent tre utiliss avec des appareils sans pices sous tension accessibles, et des connexions isoles pour 250 V. La tension de fonctionnement entre contacts relais adjacents ne doit pas dpasser 250 V. Le botier non-mtallique n'assure pas la mise la terre entre les connexions (tubes). Utiliser des traverses et des cavaliers appropris.

Instructions pour le montage


Les instructions de montage varient suivant le modle (montage mural, tuyau, rail DIN et montage panneau afficheur dport). Veuillez suivre les instructions spcifiques chaque modle.

N.B. : l'acheminement des cbles dans des conduits doit tre ralis selon les Version botier mural ou panneau

instructions fournies (page 9) avant de procder au montage du SITRANS LUT400.

Tous les modles du SITRANS LUT400 sont livrs avec une plaque de fixation arrire. La version pour montage en panneau est livre avec un afficheur (LUI) et un couvercle simple (sans afficheur).
plaque de fixation arrire

botier dot de couvercle afficheur optionnel clips (x 2) clips (x 2)

N.B. : fixations pour montage mural non fournies.

Page FR-4

SITRANS LUT400 (HART)

7ML19985XU81

plaque de fixation arrire

Franais

botier dot de couvercle simple clips (x 2) clips (x 2)

N.B. : fixations pour montage mural non fournies. partie postrieure du botier

pices rainures

Montage mural du botier


1) 2) 3) Marquer et percer les trous pour les quatre vis de montage (fournies par le client). Serrer les vis avec un tournevis. Aligner les pices rainures sur la partie postrieure du botier avec les clips sur la plaque de fixation arrire. Placer le LUT400 sur la plaque arrire, appuyer correctement et faites glisser vers le bas pour fixer l'appareil en place.

orifices pour vis de montage sur la plaque arrire

Note importante:
Taille recommande des fixations : vis M8 ou 5/16" avec rondelle, diamtre extrieur 17 mm maximum ou 5/8" Montage conseill : fixation directe au mur. Toute autre surface prvue pour le montage DOIT pouvoir supporter un poids quatre fois suprieur celui de l'appareil.

7ML19985XU81

SITRANS LUT400 (HART)

Page FR-5

Couvercle dport

Franais

couvercle afficheur dport

joint

4 fixations pour montage sur panneau fournies cble d'extension SITRANS LUT400

Montage du couvercle dport


N.B. : le couvercle pour montage dport peut tre install une distance max. de 5 m de l'appareil. Deux extensions de cble de 2,5 m sont fournies cet effet (en option). Se reporter aux Instructions de service. 1) Dcouper le panneau l'aide du gabarit fourni pour pouvoir y insrer le couvercle quip d'un afficheur LUI. Placer le joint l'intrieur du couvercle, et aligner les trous de fixation. Aligner la partie postrieure du couvercle-afficheur avec la dcoupe panneau. Marquer la surface de montage et percer les trous ncessaires pour les quatre vis de fixation (fournies). Serrer avec un tournevis et une cl. N.B. : couple recommand pour assurer une bonne tanchit des vis de fixation : 1,1 N m 10 in-lbs

2)

Remarques :
Montage conseill : montage sur panneau, cart max. 5 m de l'appareil. Toute autre surface utilise pour le montage DOIT pouvoir supporter un poids quatre fois suprieur celui de l'appareil. N.B. : fixations fournies : vis M5, rondelled d'tanchit, rondelled plated M5 et croud. Ces fixations sont requises pour garantir l'indice de protection IP65 du couvercle dport. Page FR-6 SITRANS LUT400 (HART) 7ML19985XU81

Montage sur support-tuyau

Franais

plaque de fixation arrire dispositif de blocage botier dot de couvercle afficheur optionnel

boulon en U

tuyau N.B. : fixations pour montage sur support-tuyau non fournies.

Montage du botier
1) 2) Fixer la plaque de fixation arrire au support-tuyau avec des boulons en U et des dispositifs de blocage (fourniture client) adapts au diamtre du tuyau. Serrer les vis avec un tournevis. Eviter tout serrage excessif qui pourrat provoquer la torsion de la plaque. Cette dformation aurat pour consquence d'empcher le montage du LUT400 sur la plaque de fixation arrire. Installer l'appareil sur la plaque de fixation arrire (tel que dcrit l'tape 3, Montage mural du botier page 5).

3)

Remarque importante :
Montage conseill : directement sur un support-tuyau horizontal ou vertical. Toute autre surface utilise pour le montage DOIT pouvoir supporter un poids quatre fois suprieur celui de l'appareil. Dimensions conseilles du support-tuyau : maximum : tuyau 3" minimum : tuyau 3/4" Taille recommande des fixations : Boulons en U : maximum : dimension tuyau 3" avec filetage M8 ou 3/8" minimum : dimension tuyau 3/4" avec filetage M6 ou 1/4" Ecrous hexagonaux : M6 ou 1/4" - M8 ou 3/8" Rondelle : maximum : diamtre extrieur 16 mm ou 13/16".

7ML19985XU81

SITRANS LUT400 (HART)

Page FR-7

Montage sur rail DIN


partie postrieure du botier

Franais

Rail DIN

clips (x 2)

fente

tiroir rail DIN

Montage du botier
1) 2) 3) Incliner la partie suprieure du botier vers le rail DIN ; la placer juste au dessus de l'extrmit suprieure du rail. Dplacer le botier vers le bas, contre le rail DIN pour attacher les clips situs au dos du botier, sur la partie suprieure du rail DIN. Placer l'appareil contre le rail DIN pour l'insrer dans le tiroir du rail DIN. Cette opration permet de bien fixer le LUT400 au rail DIN.

Remarque importante :
Montage conseill : directement sur le rail DIN horizontal. Dimensions requises du rail DIN : TH 35-7.5 ou TH 35-15 conformment la norme CEI 60715. Le rail DIN DOIT pourvoir supporter un poids quivalent quatre fois le poids du SITRANS LUT400.

Dmontage du botier
1) Depuis la face avant de l'appareil, placer le tournevis dans la fente situe sur la partie infrieure du tiroir du rail DIN, et pousser vers le bas. Cette opration permet de dclipser le tiroir de la partie infrieure du rail DIN. Maintenir le tiroir et place et pousser le botier vers le haut pour dclipser l'appareil de la partie suprieure du rail DIN.

2)

Page FR-8

SITRANS LUT400 (HART)

7ML19985XU81

Prparation pour l'acheminement des cbles


Les cbles peuvent tre insrs dans le botier travers un conduit ou des presse-toupes. Suivre les tapes 1 5 pour dgager les orifices d'entre des cbles. Complter ensuite les tapes ci-dessous pour acheminer les cbles dans un conduit ou des presse-toupes. 1) 2) 3) 4) 5) Maintenir le couvercle du botier ferm et les vis de fixation serres. Placer la pointe du tournevis dans la rainure situe sur la partie extrieure de l'orifice de passage (cf. illustration ci-dessous). Taper avec la paume de la main l'extrmit du tournevis pour dgager le cache de l'orifice. Desserer les vis et retirer le couvercle du botier. Retirer les pices en plastique utilises pour couvrir les entres des cble. Veiller ne pas endommager l'lectronique avec l'lectricit statique ou les outils employs pour dgager les entres de cble.
SITRANS LUT400

Franais

Notes :
Une fois dgages, les entres de cbles ont un diamtre qui varie de 21,4 mm 21,6 mm. Les presse-toupes M20 (diamtre 20 mm), et les conduits 1/2" NPT (diamtre 21,3 mm) sont adapts cet orifice. Choisir le joint d'tanchit utilis pour les entres de cbles avec prcaution. L'utilisation d'un joint face plate est fortement conseille (viter les joints toriques). En cas d'utilisation de presse-toupes diffrentes l'utilisateur est responsable de maintenir la classification IP65 des orifices d'entre.

orifices d'entre des cbles

cache de l'orifice de passage (x 3)

rainure pour tournevis (x 3)

Cble achemin travers le conduit


(suite des tapes 1 5 ci-dessus) 6) Une fois l'entre des cbles prpare (tapes 1 5 ci-dessus), fixer le conduit au botier uniquement avec des raccords de taille adapte, tanches l'eau (taille conduit 1/2" NPT.) 7) Remettre le couvercle en place et serrer les vis.

Cble expos, achemin dans les presse-toupes


(suite des tapes 1 5 ci-dessus) 6) Une fois le botier prpar pour le passage des cbles (tapes 1 5 cidessus), desserrer les presse-toupes et les fixer fermement au botier. 7) Acheminer les cbles travers les presse-toupes. Sparer le cble de l'alimentation des cbles de transmission du signal puis relier les cbles aux borniers de connexion. 8) Serrer les presse-toupes pour garantir ltanchit. 9) Fermer le couvercle et serrer les vis. Page FR-9

7ML19985XU81

SITRANS LUT400 (HART)

Raccordement lectrique
Remarques de scurit concernant le raccordement
Pour plus de dtails sur le cblage, consulter la version complte des Instructions de service. Tous les composants associs l'instrument de mesure doivent tre installs suivant les instructions fournies. Relier les blindages de cble aux borniers appropris du LUT400 (borniers identifis par sur l'appareil). Pour viter toute diffrence de potentiel la terre, veiller relier les blindages de cble la terre. Veiller limiter la longueur des fils apparents des cbles blinds pour minimiser les interfrences (missions alatoires, bruits parasites, ...).

Franais

Raccordement du SITRANS LUT400


AVERTISSEMENTS :

Consulter la plaque signaltique de lappareil pour vrifier la conformit avec les spcifications requises. Pour garantir la protection IP/NEMA veiller l'tanchit des conduits.
Notes : Des cbles et des conduits spars peuvent tre ncessaires pour garantir la conformit avec les consignes de cblage ou les normes lectriques. Pour accder au compartiment de cblage : 1) 2) 3) Dvisser les vis de blocage 1/4 de tour. Soulever le couvercle et le rabattre gauche sur les charnires. Le couvercle peut tre laiss ouvert, et scuris au moyen des charnires. Il peut aussi tre libr des charnires et laiss sur le ct, pour accder au compartiment de cblage.

4) 5) 6)

Raliser les connexions tel qu'indiqu ci-dessous. Une fois le cblage ralis, remettre le couvercle de l'appareil en place. Serrer les vis de fixation.

Page FR-10

SITRANS LUT400 (HART)

7ML19985XU81

Compartiment de cblage
La plaque bornes du LUT400 permet la connexion simultane des entres / sorties. Les borniers de connexion peuvent tre retirs pour faciliter le cblage.

Franais

AVERTISSEMENT : veiller raliser les connexions correctement au moment de rinstaller l'appareil. Les erreurs de connexion peuvent entraner l'endommagement de l'appareil ou de tout quipement externe raccord ce dernier. N.B. : couple de serrage recommand des vis de fixation du bornier. 0,56 - 0,79 N m 5 - 7 in-lbs Veiller ne pas trop serrer les vis. alimentation
TS-3 TRANSDUCER DISCRETE INPUTS INPUT POWER

alimentation lectrique
14

8 1

9 2

10

11 0.25 A, 250 V SLOW BLOW

12 L1

13 L2/N 100 - 230 V

DC SYNC

SHLD

WHT

BLK

BLK

WHT

borniers de
mA OUT/HART BATTERY BR2032

batterie

RELAYS

DISPLAY 15 16 RLY 1 17 18 19 20 21 22 23 24

RLY 2

RLY 3 ACT

PAS SERVICE

connexion USB

cble afficheur

connexion mA HART

Alimentation lectrique
AVERTISSEMENTS : Les bornes d'entre CC doivent tre alimentes par une source mme de fournir l'isolation lectrique entre/sortie requise pour la conformit avec les rgles de scurit de la norme CEI 61010-1. Isoler tous les cblages en tenant compte des tensions nominales. Note : La mise sous tension initiale du SITRANS LUT400 doit tre ralise sans actionner les appareils d'alarme ou de contrle ventuellement utiliss. Ces derniers doivent tre mis en service aprs vrification du bon fonctionnement de l'appareil de mesure principal.

7ML19985XU81

SITRANS LUT400 (HART)

Page FR-11

Le SITRANS LUT400 est propos en version CA ou CC.

Alimentation CA Franais
INPUT POWER

Alimentation CC
INPUT POWER

12

13

14

12 10 - 32 V

13

14

L1 L2/N 100 - 230 V

Courant alternatif : 100-230 V CA 15%, 50/60 Hz, 36 VA (10W)

Courant continu : 10-32 V CC, 10W

Note : veiller ce que l'appareil soit reli une prise de terre fiable. 1) 2) 3) Pour connecter l'appareil l'alimentation lectrique, retirer environ 70 mm (2.75") de gaine de l'extrmit du cble, puis insrer les cbles dans le presse-toupe 1). Connecter les cbles aux borniers tel quillustr : la polarit est indique sous le bornier de connexion. Effectuer la mise la terre suivant les dispositions rglementaires locales. Remarques concernant la connexion de lalimentation CA : L'appareil doit tre protg par un fusible 15 A ou par un disjoncteur prvu cet effet. Prvoir un disjoncteur ou un commutateur de mise hors service clairement identifi dans l'installation. Ce dernier doit tre facilement accessible et doit se trouver proximit de l'appareil.

Batterie
Le SITRANS LUT400 est fourni avec une batterie de sauvegarde. (Pour plus de dtails sur l'emplacement de la batterie, se reporter la page 11.) La batterie (type BR2032) a une dure de vie de dix ans. Son efficacit diminue en fonction de la temprature ambiante. En cas de coupure de l'alimentation lectrique, la batterie assure le fonctionnement de l'horloge temps rel du LUT400 jusqu'au retour de l'alimentation secteur (pour remplacer la batterie, se reporter la version intgrale des Instructions de service).

1)Lorsque le cble est install sous un tube protecteur, utiliser des bouchons tanches adapts.

Page FR-12

SITRANS LUT400 (HART)

7ML19985XU81

Transducteurs
Avertissement : Les bornes de connexion des transducteurs prsentent une tension dangereuse durant le fonctionnement. Le cble du transducteur doit tre install sous un conduit mtallique mis la terre, sans aucun autre cble (except celui du capteur de temprature Notes: TS-3, si utilis) Ne pas connecter le SITRANS LUT400 avec un cble coaxial compte tenu de la haute tension transmise sur le blindage. Il est impratif de connecter le blindage du LUT400 et le cble blanc du transducteur sparment, des borniers diffrents. Ne pas tenir compte des consignes contradictoires fournies dans les versions
TS-3 TRANSDUCER

Franais

SHLD

BLK

WHT

Relais
Les relais sont illustrs en tat dsactiv. Tous les relais peuvent fonctionner en logique positive ou ngative (cf. paramtre 2.8.11.Relay Logic (Logique relais), dans la version intgrale des Instructions de service).
RELAYS

BLK

WHT

Coupure de courant
Relais 2, 3 normalement ouverts. Relais 1 normalement ouvert ou normalement ferm. En cas de perte de l'alimentation lectrique, les relais basculent dans leur tat normal d'origine.

15

16 RLY 1

17

18

19

20

21

RLY 2

RLY 3

Types de relais
Un relais inverseur SPDT (NO ou NC) (relais 1), 1A 250 V CA, charge ohmique, 3A 30 V CC Deux relais NO (relais 2, 3), 5A 250 V CAC, charge ohmique, 3A 30 V CC

7ML19985XU81

SITRANS LUT400 (HART)

Page FR-13

Connexion HART Sortie analogique (HART) Franais


mA OUT/HART

Pour une connexion HART ACTIVE (avec l'alimentation lectrique intgre du LUT400), connecter les bornes 22 et 23.

22

23

24

Pour une connexion HART PASSIVE (avec une alimentation lectrique externe), connecter les bornes 23 et 24.

ACT

PAS

Pour plus de dtails, se reporter aux paramtres de sortie analogique (2.5.Current Output, Sortie analogique) dans la section Paramtres (Instructions de service, version intgrale).

Connexion en zone dangereuse


Cf. page 35.

Mise en service du SITRANS LUT400


Le LUT400 doit, de prfrence, tre configur en suivant ces tapes : lancer d'abord l'Assistant de mise en service correspondant votre application (niveau, volume, dbit). Rgler ensuite les pompes avec l'assistant de contrle de pompage (si applicable). Enfin, configurer les alarmes ou autres contrles, totalisateurs et chantillonneurs l'aide des paramtres correspondants (se reporter la version intgrale des Instructions de service). Les Assistants de mise en service sont accessibles : localement (cf. Mise en service rapide avec l'interface pour l'utilisateur (LUI) page 17) ou distance (cf. Assistants de mise en service complmentaires : page 17)

Activation du SITRANS LUT400


Mettre l'instrument sous tension. Le SITRANS LUT400 dmarre automatiquement en mode de Mesure . Un cran de transition indiquant le logo Siemens, suivi de la rvision du micrologiciel LUI est affich pendant l'laboration de la premire mesure. Le choix de la langue d'interface (anglais, allemand, franais, espagnol ou chinois) est propos lors de la configuration initiale de l'appareil. Avant de configurer l'appareil, veiller rgler la date et l'heure. (Le paramtre 2.14.Date and Time (Date et heure) permet les rglages ncessaires. Pour plus de dtails se reporter la version intgrale des Instructions de service.)

Page FR-14

SITRANS LUT400 (HART)

7ML19985XU81

Affichage LCD Affichage en mode de mesure : Fonctionnement normal 1) Franais Niveau


1 10
SITRANS LUT400 NIVEAU

2
1/2

4 5

18.91
1

6
2

5 SAUVEGUARDE PARAMTRES

Dbit
1 10 11 12
SITRANS LUT400 DBIT HAUTEUR...
1 2 3

3
1/2
L/S M

4 5 6

500313..
60.000

TOTALISATE...

999999.99

5 SAUVEGUARDE PARAMTRES

1 marqueur 2 valeur mesure (niveau, espace, distance, volume, dbit, ou hauteur de lame) 3 valeur affiche [Variable primaire (PV)=1 de 2, Variable secondaire (SV)=2 de 2] 4 indicateur d'affichage altern1) pour PV ou SV 5 units 6 reprsentation du niveau (bargraphe) 7 zone d'affichage secondaire, indique les relais configurs (gauche) et les entres TOR (droite) 8 zone ddie laffichage de messages de texte 9 indicateur de ltat de fonctionnement de l'appareil 10 mode slectionn (mode principal) : niveau, espace, distance, volume, hauteur de lame ou dbit 11 mode secondaire = hauteur de lame lorsque le mode principal = dbit 12 valeurs totalisateur : affichage altern : totalisateur journalier et totalisateur en service

Dfaut signal
0 LOE

8 zone de texte ; indication d'un code d'erreur/message d'erreur 9 icne maintenance requise

Affichage en mode Program Visualisation durant la navigation

actuel / La barre de menu apparat si la numro TRANSDUCTEUR nom de 2.1.6 longueur du menu dpasse la de paraPAS DE CAPT paraXRS-5 mtre capacit d'affichage de l'cran. mtre XPS-10 bande La largeur de la bande item dans la XPS-15 item XCT-8 barre de menu reflte la longueur du d'item actuel XCT-12 XPS-30 menu : plus la bande est large, plus le barre de menu nombre d'items est rduit. La position de la bande item reflte la position approximative de l'item actuel dans la liste. Si la bande est situe au milieu de la barre de menu, l'item actuel se trouve au milieu de la liste. Plus la bande est large, plus le nombre d'items est rduit.

menu

1)Appuyer sur

ou

pour commuter l'affichage.

7ML19985XU81

SITRANS LUT400 (HART)

Page FR-15

Visualisation de paramtres Franais


menu associ au paramtre nom de paramtre
CAPTEUR
UNITS CAPTEUR MODE CAPTEUR MODE CAPTEUR SECONDAIRE

Visualisation en mode dition


2.1.6

numro de paramtre valeur/ slection de paramtre

TRANSDUCTEUR
PAS DE CAPT XPS-10 XPS-15 XCT-8 XCT-12 XPS-30

2.1.6

XRS-5

TRANSDUCTEUR

FRQUENCE

XRS-5

Mode PROGRAM
Utiliser les boutons poussoirs intgrs : appuyer sur accder au menu niveau 1. Parcourir le menu l'aide de pour accder au mode PROGRAM et , , , .

Slection d'une option


a) Accder au paramtre souhait. b) Appuyer sur la flche DROITE slectionne est surligne. pour accder au mode Edition. La valeur

c) Accder la nouvelle valeur slectionne. d) Appuyer sur la flche DROITE pour valider. L'affichage LCD revient aux paramtres et indique la nouvelle valeur slectionne.

Modification d'une valeur numrique


a) Accder au paramtre souhait. Par exemple : Point d'talonnage min. (2.2.1.). b) Une fois le paramtre slectionn, sa valeur actuelle est affiche. c) Appuyer sur la flche DROITE pour accder au mode Edition. La position du curseur est surligne. d) Utiliser les flches GAUCHE et DROITE pour dplacer le curseur sur le chiffre modifier. e) Au fur et mesure que chaque chiffre est slectionn, utiliser les flches HAUT et BAS f) pour augmenter ou diminuer la valeur correspondante. et BAS
PT DTAL. MIN. 2.2.1

ESC

+ _ _ _ _ _ _ 2. 0 M

OK

Lorsque la virgule dcimale est slectionne, utiliser les flches HAUT pour la dplacer.

g) Pour quitter le mode dition sans sauvegarder les modifications, appuyer sur la flche GAUCHE en continu jusqu' ce que ESC soit slectionn. Appuyer de pour quitter sans sauvegarder. Autrement, si la en nouveau sur la flche GAUCHE

nouvelle valeur du paramtre est correcte, appuyer sur la flche DROITE continu jusqu' ce que OK soit slectionn.

h) Appuyer sur la flche DROITE pour confirmer la nouvelle valeur. L'affichage LCD revient aux paramtres et indique la nouvelle valeur slectionne. Vrifier l'exactitude des valeurs. Page FR-16 SITRANS LUT400 (HART) 7ML19985XU81

Assistants de mise en service


La configuration du SITRANS LUT400 pour la mesure de niveau, de volume (gomtries de cuve standards), ou de dbit peut tre ralise avec la Mise en service rapide avec l'interface pour l'utilisateur (LUI) dcrite ci-dessous. Un Assistant ddi pour les applications avec des cuves gomtrie complexe est disponible via SIMATIC PDM. Se reporter Assistants de mise en service via SIMATIC PDM page 33.

Franais

Assistants de mise en service complmentaires :


D'autres assistants de mise en service sont disponibles travers d'autres logiciels 1): SIMATIC PDM (HART) AMS (HART) FC375/475 (HART) FDTs (HART) Des tableaux de configuration des paramtres (permettant de noter les valeurs avant de configurer l'appareil l'aide d'un Assistant de mise en service) sont disponibles sur notre site internet. Consulter www.siemens.com/sitransLUT400 > Support > Application Guides.

Mise en service rapide avec l'interface pour l'utilisateur (LUI)


1) 2) Appuyer sur pour accder au mode Programmation. Slectionner Assistants (1.), Mise en service rapide (1.1.), et le mode de mise en service souhait : Mode Niveau (1.1.1.), Mode Volume (1.1.2.), ou Mode Dbit (1.1.3.). [Sur les modles configurs, affichage de l'Assistant de mise en service Dbit sur les versions LUT430 (Pompage et Dbit), et LUT440 (Dbit en canal ouvert).] Suivre les tapes indiques puis slectionner Finish (Terminer) pour sauvegarder les paramtres rgls et revenir au menu mode Programmation. Appuyer trois fois sur pour revenir au mode de Mesure. Notes : Les rglages raliss l'aide des Assistants de mise en service rapide sont interdpendants. Par consquent, les modifications prennent effet uniquement aprs avoir slectionn Finish (Terminer) la fin de la procdure. Les Assistants de mise en service ne doivent pas tre utiliss pour modifier les paramtres individuellement : pour plus de dtails, se reporter la section Paramtres, dans la version intgrale des Instructions de service. (les rglages personnaliss doivent tre raliss aprs la procdure de mise en service rapide). Pour plus de dtails sur les paramtres accessibles durant la mise en service rapide, se reporter aux Instructions de service.

3)

1. Assistants 1.1. Mise en service rapide 1.1.1. Mode Niveau


Cet Assistant permet de configurer l'appareil pour des applications de niveau simples. a) Appuyer sur la flche DROITE pour lancer le mode PROGRAM et accder au menu niveau 1 : MENU PRINCIPAL. b) Appuyer deux fois sur la flche DROITE c) Appuyer sur la flche DROITE pour accder l'item menu 1.1.1.

pour accder la mise en service rapide-Niveau.

1)Se reporter aux instructions relatives la communication SITRANS LUT400 (HART).

7ML19985XU81

SITRANS LUT400 (HART)

Page FR-17

d) A chaque tape, appuyer sur la flche BAS pour confirmer les valeurs par dfaut et accder directement l'item suivant,

MISE EN SERVICE RAPIDE NIVEAU


TRANSDUCTEUR XRS-5
RETOUR ANNULER SUIVANT EDITER

Franais

ou sur la flche DROITE pour accder au mode Edition : la valeur slectionne est surligne. e) Accder l'item souhait et appuyer sur flche DROITE pour sauvegarder la modification. Appuyer sur flche BAS pour continuer. f) A tout moment, vous pouvez appuyer sur la flche HAUT ou sur la flche GAUCHE pour quitter l'Assistant. Dmarrage de l'Assistant de mise en service-Niveau

MISE EN SERVICE RAPIDE NIVEAU


TRANSDUCTEUR

PAS DE CAPT XPS-10 XPS-15 XCT-8 XCT-12

XRS-5

pour revenir en arrire,

Indique le type d'Assistant qui sera lanc.

Options
Transducteur

CANCEL (ANNULER), START (DEMARRER)

Dfinit le modle de transducteur Siemens connect l'appareil..

Options
Fonctionnement

NO TRANSDUCER (SANS TRANSDUCTEUR), XRS-5, XPS-10, XPS-15, XCT-8, XCT-12, XPS-30, XPS-40, XLT-30, XLT-60, STH Valeur par dfaut : NO TRANSDUCER (SANS TRANSDUCTEUR)

Dfinit le type de mesure (et la sortie analogique correspondante) requis pour lapplication.
point de rfrence du capteur distance point d'talonnage max. espace

niveau point d'talonnage min.

porte max.

Mode
LEVEL (NIVEAU) SPACE (ESPACE) DISTANCE

Description
* Hauteur du produit Distance la surface du produit.

Point de rfrence
Point d'talonnage min. (niveau process vide) Point d'talonnage max. (niveau process plein) Point de rfrence du capteur (cf. page suivante)

Page FR-18

SITRANS LUT400 (HART)

7ML19985XU81

(suite de la page prcdente)

Mode

Description
NE PAS slectionner. Si le mode de Fonctionnement affich est AUTRE, l'appareil fonctionne en tant que contrleur de niveau. Nanmois le rglage prcdent ne correspondait pas aux modes LEVEL (NIVEAU), SPACE (ESPACE) ou DISTANCE. Pour lancer l'Assistant de mise en service Niveau, le mode de fonctionnement de l'appareil doit tre rgl sur NIVEAU, ESPACE, ou DISTANCE.

Point de rfrence

Franais

OTHER (AUTRE)

Source de temprature

Source de la mesure de temprature utilise pour rgler la vitesse du son..

Options
Temprature fixe

TRANSDUCER (TRANSDUCTEUR), FIXED TEMPERATURE (TEMP . FIXE), EXTERNAL TS-3 (TS-3 EXTERNE), AVERAGE OF SENSORS (MOYENNE DES CAPTEURS) Valeur par dfaut : TRANSDUCER (TRANSDUCTEUR)

Utiliser cette fonction lorsquun capteur de temprature nest pas utilis.

Valeur

Plage : -100,0 +150,0 C Valeur par dfaut : +20,0 C

Ce paramtre est affich uniquement lorsque FIXED TEMPERATURE (TEMP . FIXE) est slectionne pour Source de temprature. Units

Units de mesure capteur.

Options

M, CM, MM, FT, IN Valeur par dfaut : M

Note : dans cet exemple, toutes les valeurs sont en mtres (m).
Point d'talonnage max.

Distance entre le point de rfrence du capteur et le point d'talonnage max. : gnralement le niveau process plein.

Valeur

Plage : 0,000 60,000 Valeur par dfaut : 0,000

Point d'talonnage min.

Distance entre le point de rfrence du capteur et le point d'talonnage min. : gnralement le niveau process vide.

Valeur

Plage : 0,000 60,000 Valeur par dfaut : 60,000

7ML19985XU81

SITRANS LUT400 (HART)

Page FR-19

Dfinit la vitesse de raction de l'appareil aux variations de mesure dans la plage spcifie.

Temps de rponse

Franais

N.B. :

Le rglage du Temps de rponse s'effectue uniquement via l'Assistant de mise en service. Toute modification des paramtres Fill Rate per Minute (2.3.1.) (Vitesse de remplissage par min.), Empty Rate per Minute (2.3.2.) (Vitesse de vidange par min.), ou Damping Filter (2.3.3.) (filtre damortissement) aprs la fin de l'assistant annule et remplace le Temps de rponse rgl auparavant. Temps de rponse est toujours indiqu en m/minute.
SLOW (LENT) (0,1 M/MIN) MEDIUM (MOYEN) (1,0 M/MIN) FAST (RAPIDE) (10 M/MIN) Valeur par dfaut : SLOW (LENT) (0,1 M/MIN)

Options

Slectionner une vitesse lgrement suprieure la vitesse de remplissage ou de vidange maximale (la plus leve des deux). Plus la rponse est lente, plus la prcision sera leve. Inversement, plus la rponse est rapide, plus les variations de niveau seront importantes. Fin de l'Assistant de mise en service-Niveau

La bonne excution de la procdure de mise en service suppose la prise en compte des modifications.

Options

BACK (RETOUR), CANCEL (ANNULER), FINISH (TERMINER) (Le menu de mise en service rapide 1.1 est affich ds la fin de l'assistant (ou lorsque celui-ci est annul). Lorsque CANCEL (ANNULER) est slectionn, aucune valeur n'est enregistre.)

Pour transfrer les valeurs rgles durant la mise en service vers l'appareil et revenir (Terminer). Appuyer trois fois sur au mode Programmation, utiliser la flche BAS la flche GAUCHE pour revenir au mode de Mesure.

1.1.2. Mode Volume

Cet assistant permet de configurer l'appareil pour la mesure de volume dans des cuves de gomtrie standard.
a) b) c) d) Appuyer sur la flche DROITE pour lancer le mode PROGRAM et accder au menu niveau 1 : MAIN MENU (MENU PRINCIPAL). Appuyer deux fois sur la flche DROITE Appuyer sur la flche BAS rapide-Volume. pour accder l'item menu 1.1.1. pour accder la mise en service
MISE EN SERVICE RAPIDE VOLUME
TRANSDUCTEUR XRS-5
RETOUR ANNULER SUIVANT EDITER

et sur la flche DROITE

A chaque tape, appuyer sur la flche BAS pour confirmer les valeurs par dfaut et accder directement l'item suivant, ou sur la flche DROITE pour accder au mode Edition : la valeur slectionne est surligne.

e)

Accder l'item souhait et appuyer sur flche DROITE pour sauvegarder la modification. Appuyer sur flche BAS pour continuer.

MISE EN SERVICE RAPIDE VOLUME


TRANSDUCTEUR

PAS DE CAPT XPS-10 XPS-15 XCT-8 XCT-12

XRS-5

f)

A tout moment, vous pouvez appuyer sur la flche HAUT pour revenir en arrire, ou sur la flche GAUCHE quitter l'Assistant. pour

Page FR-20

SITRANS LUT400 (HART)

7ML19985XU81

Dmarrage de l'Assistant de mise en service-Volume

Indique le type d'Assistant qui sera lanc.

Options
Transducteur

CANCEL (ANNULER), START (DEMARRER)

Franais

Dfinit le modle de transducteur Siemens connect l'appareil..

Options

NO TRANSDUCER (SANS TRANSDUCTEUR), XRS-5, XPS-10, XPS-15, XCT8, XCT-12, XPS-30, XPS-40, XLT-30, XLT-60, STH Valeur par dfaut : NO TRANSDUCER (SANS TRANSDUCTEUR)

Source de temprature

Source de temprature utilise pour rgler la vitesse du son.

Options
Temprature fixe

TRANSDUCER (TRANSDUCTEUR), FIXED TEMPERATURE (TEMP . FIXE), EXTERNAL TS-3 (TS-3 EXTERNE), AVERAGE OF SENSORS (MOYENNE DES CAPTEURS) Valeur par dfaut : TRANSDUCER (TRANSDUCTEUR)

Utiliser cette fonction lorsquun capteur de temprature nest pas utilis.

Valeur

Plage : -100,0 +150,0 C Valeur par dfaut : +20,0 C

Ce paramtre est affich uniquement lorsque FIXED TEMPERATURE (TEMP . FIXE) est slectionne pour Source de temprature. Forme de cuve

Dfinit la forme de la cuve, permettant au SITRANS LUT400 de calculer le volume (et non le niveau). Lorsque NONE est slectionn il n'y a pas de conversion de volume. Slectionner une forme qui correspond celle de la cuve/rservoir contrl(e).
NONE (AUCUN), LINEAR (LINEAIRE), CYLINDER (CYLINDRE), PARABOLIC BOTTOM (FOND PARABOLIQUE), HALF SPHERE BOTTOM (FOND HEMISPHERIQUE), FLAT SLOPED BOTTOM (FOND PLAT INCLINE), PARABOLIC ENDS (EXTREM.PARABOLIQUES), SPHERE (SPHERE), CONICAL BOTTOM (FOND CONIQUE), CURVE TABLE (TABLE COURBE), LINEAR TABLE (TABLE LINEAIRE) Valeur par dfaut : LINEAR (LINEAIRE)

Options

Lorsque CURVE TABLE (TABLE COURBE) ou LINEAR TABLE (TABLE LINEAIRE) est slectionn, entrer les valeurs correspondantes aux points de contrle niveau et volume la fin de l'assistant de mise en service. (Pour plus de dtails se reporter au Paramtre 2.6.7.Table 1-8, dans les Instructions de service.) Units

Units de mesure capteur.

Options

M, CM, MM, FT, IN Valeur par dfaut : M

N.B. : dans cet exemple, toutes les valeurs sont en mtres (m).

7ML19985XU81

SITRANS LUT400 (HART)

Page FR-21

Point d'talonnage max.

Distance entre le point de rfrence du capteur et le point d'talonnage max. : gnralement le niveau process plein.

Franais

Valeur

Plage : 0,000 60,000 Valeur par dfaut : 0,000

Point d'talonnage min.

Distance entre le point de rfrence du capteur et le point d'talonnage min. : gnralement le niveau process vide.

Valeur

Plage : 0,000 60,000 Valeur par dfaut : 60,000

Temps de rponse

Dfinit la vitesse de raction de l'instrument aux variations de mesure dans la plage spcifie.

N.B. :

Le rglage du Temps de rponse s'effectue uniquement via l'Assistant de mise en service. Toute modification des paramtres Fill Rate per Minute (2.3.1.) (Vitesse de remplissage par min.) ou Empty Rate per Minute (2.3.2.) (Vitesse de vidange par min.) aprs la fin de l'assistant annule et remplace le Temps de rponse rgl auparavant. Temps de rponse est toujours indiqu en m/minute.
SLOW (LENT) (0,1 M/MIN) MEDIUM (MOYEN) (1,0 M/MIN) FAST (RAPIDE) (10 M/MIN) Valeur par dfaut : SLOW (LENT) (0,1 M/MIN)

Options

Slectionner une vitesse lgrement suprieure la vitesse de remplissage ou de vidange maximale (la plus leve des deux). Plus la rponse est lente, plus la prcision sera leve. Inversement, plus la rponse est rapide, plus les variations de niveau seront importantes. Dimension A

Hauteur du fond de la cuve lorsque le fond de la cuve est conique, pyramidal, parabolique, sphrique ou plat inclin..

Valeur
Dimension L

Plage : 0,000 99.999 Valeur par dfaut : 0,000

Longueur de la section cylindrique d'une cuve horizontale avec extrmits paraboliques.

Valeur
Units de volume

Plage : 0,000 99,999 Valeur par dfaut : 0,000

Dfinit le volume en units de mesure.

Options

L, USGAL, IMPGAL, CUM, USER DEFINED * (DEFINI PAR L'UTILISATEUR) Valeur par dfaut : L

* Lorsque l'option USER DEFINED (DEFINI PAR L'UTILISATEUR) est slectionne, la valeur doit tre rgle la fin de l'assistant de mise en service. Pour plus de dtails se reporter au Paramtre 2.6.6.User Defined Unit (Unit dfinie par l'utilisateur) (Instructions de service). Page FR-22 SITRANS LUT400 (HART) 7ML19985XU81

Volume maximum

Volume maximum de la cuve. Entrer le volume de la cuve correspondant au Point d'talonnage max. Par exemple, si le volume maximum de la cuve est 8000 L, entrer 8000.

Franais

Valeur

Plage : 0,0 9999999 Valeur par dfaut : 100,0

Fin de l'Assistant de mise en service-Volume

La bonne excution de la procdure de mise en service suppose la prise en compte des modifications.

Options

BACK (RETOUR), CANCEL (ANNULER), FINISH (TERMINER) (Le menu de mise en service rapide 1.1 est affich ds la fin de l'assistant (ou lorsque celui-ci est annul). Lorsque CANCEL (ANNULER) est slectionn, aucune valeur n'est enregistre.)

Pour transfrer les valeurs rgles durant la mise en service vers l'appareil et revenir au mode Programmation, utiliser la flche BAS la flche GAUCHE (Terminer). Appuyer trois fois sur pour revenir au mode de Mesure.

1.1.3. Mode Dbit


Cet Assistant permet de configurer l'appareil pour des applications de mesure de dbit simples. [Visible uniquement sur les modles configurs, versions LUT430 (Pompage et Dbit), et LUT440 (Dbit en canal ouvert).] a) Appuyer sur la flche DROITE pour lancer le mode PROGRAM et accder au menu niveau 1 : MAIN MENU (MENU PRINCIPAL). b) Appuyer deux fois sur la flche DROITE c) Appuyer deux fois sur la flche BAS la mise en service rapide-Dbit. pour accder l'item menu 1.1.1. et sur la flche DROITE pour accder

d) A chaque tape, appuyer sur la flche BAS pour confirmer les valeurs par dfaut et accder directement l'item suivant, ou sur la flche DROITE pour accder au mode Edition : la valeur slectionne est surligne. e) Accder l'item souhait et appuyer sur flche DROITE pour sauvegarder la modification. Appuyer sur flche BAS pour continuer. f)

ASSISTANT MISE EN SCE DEBIT


TRANSDUCTEUR XRS-5
RETOUR ANNULER SUIVANT EDITER

ASSISTANT MISE EN SCE DEBIT


TRANSDUCTEUR

PAS DE CAPT XPS-10 XPS-15 XCT-8 XCT-12

XRS-5

A tout moment, vous pouvez appuyer sur la flche HAUT ou sur la flche GAUCHE pour quitter l'Assistant. Dmarrage de l'Assistant de mise en service-Dbit

pour revenir en arrire,

Indique le type d'Assistant qui sera lanc.

Options
7ML19985XU81

CANCEL (ANNULER), START (DEMARRER)

SITRANS LUT400 (HART)

Page FR-23

Transducteur

Dfinit le modle de transducteur Siemens connect l'appareil..

Franais

Options

NO TRANSDUCER, XRS-5, XPS-10, XPS-15, XCT-8, XCT-12, XPS-30, XPS-40, XLT-30, XLT-60, STH Valeur par dfaut : NO TRANSDUCER (SANS TRANSDUCTEUR)

Source de temprature

Source de temprature utilise pour rgler la vitesse du son.

Options
Temprature fixe

TRANSDUCER, FIXED TEMPERATURE (TEMP .FIXE), EXTERNAL TS-3 (TS-3 EXTERNE), AVERAGE OF SENSORS (MOYENNE) Valeur par dfaut : TRANSDUCER (TRANSDUCTEUR)

Source de temprature utilise pour rgler la vitesse du son.

Valeur

Plage : -100,0 +150,0 C Valeur par dfaut : +20,0 C

Ce paramtre est affich uniquement lorsque FIXED TEMPERATURE (TEMP . FIXE) est slectionne pour Source de temprature. Dispositif de mesure primaire

Dfinit le dispositif de mesure primaire (DMP) utilis dans l'application.


EXPONENTIAL DEVICES, RECTANGULAR FLUME BS-3680 (CANAL RECT. BS-3680), ROUND NOSE HORIZONTAL CR. (DEVERSOIR HORIZONTAL A SEUIL EPAIS ARRONDI) BS-3680, TRAPEZOIDAL FLUME BS-3680 (CANAL TRAPEZOIDAL BS-3680), U-FLUME BS-3680 (CANAL A COL EN U BS-3680), FINITE CREST WEIR BS-3680 (DEVERSOIR RECTANGULAIRE A SEUIL EPAIS BS-3680), THIN PLATE RECT. (DEVERSOIR A ECHANCRURE RECTANGULAIRE) WEIR BS-3680 (DEVERSOIR BS-3680), THIN PLATE VNOTCH WEIR BS-3680 (DEVERSOIR RECT. A ECHANCRURE TRIANGULAIRE EN MINCE PAROI BS-3680) WEIR CONTRACTED (DEV. CONTRACTE), ROUND PIPE (CANALISATION CIRCULAIRE), PALMER BOWLUS FLUME (CANAL PALMER BOWLUS), H-FLUME (CANAL H), OTHER (AUTRES)* Valeur par dfaut : EXPONENTIAL DEVICES (SYSTEMES EXPONENTIELS)

Options

* L'option est rgle sur OTHER (AUTRE) en cas d'utilisation pralable de l'Assistant via un outil logiciel HART (par ex. SIMATIC PDM), et le rglage de l'appareil sur OFF ou sur UNIVERSAL HEAD VS. FLOW (HAUTEUR DE LAME UNIVERSELLE VS. DEBIT).Lors de la premire configuration le dispositif de mesure primaire (DMP) peut uniquement tre rgl sur OFF (pas de calcul)), our sur linarisation (UNIVERSAL HEAD VS. FLOW (HAUTEUR DE LAME UNIVERSELLE VS. DEBIT)) via les outils logiciels compatibles HART (SIMATIC PDM, AMS, FC375/475). Se reporter aux Instructions de service.

Mthode de calcul de dbit

Dfinit la mthode utilise pour calculer le dbit.

Options
Units

ABSOLUTE (ABSOLUE), RATIOMETRIC (RATIOMETRIQUE) Valeur par dfaut : ABSOLUTE (ABSOLUE)

Units de mesure capteur.

Options

M, CM, MM, FT, IN Valeur par dfaut : M

N.B. : Dans cet exemple, toutes les valeurs sont en mtres (m).
Page FR-24 SITRANS LUT400 (HART) 7ML19985XU81

Point d'talonnage max.

Distance entre le point de rfrence du capteur et le point d'talonnage max. : gnralement le niveau process plein.

Valeur

Plage : 0,000 60,000 Valeur par dfaut : 0,000

Franais

Point d'talonnage min.

Distance entre le point de rfrence du capteur et le point d'talonnage min. : gnralement le niveau process vide.

Valeur

Plage : 0,000 60,000 Valeur par dfaut : 60,000

Temps de rponse

Dfinit la vitesse de raction de l'instrument aux variations de mesure dans la plage spcifie.

N.B. :

Le rglage du Temps de rponse s'effectue uniquement via l'Assistant de mise en service. Toute modification des paramtres Fill Rate per Minute (2.3.1.) (Vitesse de remplissage par min.) ou Empty Rate per Minute (2.3.2.) (Vitesse de vidange par min.) aprs la fin de l'assistant annule et remplace le Temps de rponse rgl auparavant. Temps de rponse est toujours indiqu en m/minute.
SLOW (LENT) (0,1 M/MIN) MEDIUM (MOYEN) (1,0 M/MIN) FAST (RAPIDE) (10 M/MIN) Valeur par dfaut : SLOW (LENT) (0,1 M/MIN)

Options

Slectionner une vitesse lgrement suprieure la vitesse de remplissage ou de vidange maximale (la plus leve des deux). Plus la rponse est lente, plus la prcision sera leve. Inversement, plus la rponse est rapide, plus les variations de niveau seront importantes. Facteurs de calcul :

N.B. :

L'affichage des cinq paramtres suivants durant la procdure de mise en service rapide est fonction du DMP choisi ci-dessus. Ces paramtres sont utiliss dans le calcul de dbit (pour plus de dtails, se reporter la version intgrale des Instructions de service). Exposant dbit (DMP = SYSTEMES EXPONENTIELS)

Exposant pour la formule de calcul de dbit.

Valeur

Plage : -999,000 9999,000 Valeur par dfaut : 1,550

7ML19985XU81

SITRANS LUT400 (HART)

Page FR-25

Facteur k (DMP = SYSTEMES EXPONENTIELS)

Franais

Constante utilise dans le calcul de dbit pour le calcul absolu (dispositif exponentiel uniquement).

Valeur
Dversoir en V

Plage : -999,000 9999,000 Valeur par dfaut : 1,000

(PMD = THIN PLATE V-NOTCH WEIR (DMP = DEVERSOIR A ECHANCRURE TRIANGULAIRE EN MINCE PAROI))

Angle de l'chancrure triangulaire utilis dans la formule de calcul de dbit.

Valeur
Pente

Plage : 25,000 95,000 Valeur par dfaut : 25,000

(PMD = TRAPEZOIDAL FLUME or ROUND PIPE (DMP = CANAL TRAPEZOIDAL ou CANALISATION CIRCULAIRE))

Pente de dbit utilise dans la formule de calcul de dbit.

Valeur

Plage : -999,000 9999,000 Valeur par dfaut : 0,000

Coefficient de rugosit (PMD = ROUND PIPE (DMP = CANALISATION CIRCULAIRE))

Coefficient de rugosit utilis dans la formule de calcul de dbit.

Valeur

Plage : -999,000 9999,000 Valeur par dfaut : 0,000

Page FR-26

SITRANS LUT400 (HART)

7ML19985XU81

Dimensions du dispositif de mesure primaire

N.B. :

Pour chaque DMP l'exception des Systmes exponentiels, et Autres, il est ncessaire de rgler jusqu quatre valeurs dimensionnelles. Lors de la procdure de mise en service rapide, l'utilisateur spcifie les cotes relatives au dispositif primaire choisi. Le libell de la valeur dimensionnelle du dispositif primaire correspondant sera affich. DMP slectionn Libell valeur dimensionnelle (Assistant) (rf. menu param) Canal rectangulaire BS-3680
APPROACH WIDTH B (LARG. B CANAL D'APPROCHE (2.15.4.5. OCM Dimension 1)) THROAT WIDTH B (LARG. B CONTRACTION (2.15.4.6. OCM Dimension 2)) HUMP HEIGHT P (HAUT. SURELEV. RADIER P (2.15.4.7. OCM Dimension 3)) THROAT LENGTH L (LONGUEUR DE LA CONTRACTION L (2.15.4.8. OCM Dimension 4))

Franais

Dversoir horizontal seuil pais arrondi BS-3680


CREST WIDTH (LARGEUR SEUIL B (2.15.4.5. OCM Dimension 1)) CREST HEIGHT (HAUTEUR SEUIL P (2.15.4.6. OCM Dimension 2)) CREST LENGTH (LONG. SEUIL L (2.15.4.7. OCM Dimension 3))

Canal trapzodal BS-3680


APPROACH WIDTH B (LARG. B CANAL D'APPROCHE (2.15.4.5. OCM Dimension 1)) THROAT WIDTH (LARG. B DE LA CONTRACTION (2.15.4.6. OCM Dimension 2)) HUMP HEIGHT (HAUT. SURELEV. RADIER P (2.15.4.7. OCM Dimension 3)) THROAT LENGTH L (LONG. CONTRACTION L (2.15.4.8. OCM Dimension 4))

Canal en U BS-3680
APPROACH DIAMETER DA (DIAM. CANAL D'APPROCHE DA (2.15.4.5. OCM Dimension 1)) THROAT DIAMETER D (DIAM. CONTRACT. D (2.15.4.6. OCM Dimension 2)) HUMP HEIGHT P (HAUT. SURELEV. RADIER P (2.15.4.7. OCM Dimension 3)) THROAT LENGTH (LONG. CONTRACTION L (2.15.4.8. OCM Dimension 4))

Dversoir rectangulaire seuil pais BS-3680


CREST WIDTH B (LARG. SEUIL B (2.15.4.5. OCM Dimension 1)) CREST HEIGHT P (HAUT. SEUIL P (2.15.4.6. OCM Dimension 2)) CREST LENGTH L (LONG. SEUIL L (2.15.4.7. OCM Dimension 3))

Dversoir chancrure rectangulaire, en mince paroi BS-3680


APPROACH WIDTH B (LARGEUR B CANAL D'APPROCHE (2.15.4.5. OCM Dimension 1)) CREST WIDTH B (LARGEUR DU SEUIL B (2.15.4.6. OCM Dimension 2)) CREST HEIGHT P (HAUTEUR DU SEUIL P (2.15.4.7. OCM Dimension 3))

Dversoir rectangulaire contract


CREST WIDTH B (LARGEUR DU SEUIL B (2.15.4.5. OCM Dimension 1))

(cf. page suivante)

7ML19985XU81

SITRANS LUT400 (HART)

Page FR-27

DMP slectionn Libell de la valeur dimensionnelle (Assistant) (rf. menu (suite) Param.) Canalisation circulaire

Franais

PIPE INSIDE DIAMETER D (DIAMETRE INT. TUYAU D (2.15.4.5. OCM Dimension 1))

Canal Palmer Bowlus


MAXIMUM FLUME WIDTH HMAX (LARGEUR MAX. DU DEVERSOIR HMAX (2.15.4.5. OCM Dimension 1))

Canal H
MAXIMUM LISTED HEAD HMAX (HAUTEUR DE LAME MAX. HMAX (2.15.4.5. OCM Dimension 1))

Hauteur de lame maximale

Valeur de niveau max. associe au DMP.

Valeur

Plage : 0,000 60,000 Valeur par dfaut : 60,000

Dcalage hauteur de lame zro

L'cart (positif ou ngatif) entre Point d'talonnage min. et la hauteur de lame zro (niveau si dbit zro).

Valeur

Plage : -60,000 60,000 Valeur par dfaut : 0,000

Units de dbit instantan

Unit de volume utilise pour afficher le dbit total.

Options

L/S, L/MIN, CUFT/S, CUFT/D, GAL/MIN, GAL/D, IMPGAL/MIN, IMPGAL/D, CUM/H, CUM/D, USER DEFINED (DEFINI(E) PAR L'UTILISATEUR) * Valeur par dfaut : L/S

* Lorsque l'option USER DEFINED (DEFINI PAR L'UTILISATEUR) est slectionne, la valeur doit tre rgle la fin de l'assistant de mise en service. Pour plus de dtails se reporter au Paramtre 2.6.6.User Defined Unit (Unit dfinie par l'utilisateur) dans les Instructions de service. Dbit maximum 20 mA

Dbit maximum.

Valeur

Plage : -999 9999999 Valeur par dfaut : 100

Dcimales pour dbit instantan

Nombre maximum de chiffres affich aprs la virgule.

Options
Dbit inhib

NO DIGITS (0 CHIFFRES), 1 DIGIT, 2 DIGITS, 3 DIGITS Valeur par dfaut : NO DIGITS (0 CHIFFRES)

Dsactive la totalisation pour les hauteurs de lame situes sur ou en dessous du seuil dfini.

Valeur
Page FR-28

Plage : 0,000 60,000 Valeur par dfaut : 0,000

SITRANS LUT400 (HART)

7ML19985XU81

Fin de l'assistant de mise en sce rapide Dbit.

La bonne excution de la procdure de mise en service suppose la prise en compte des modifications.

Franais

Options

BACK (RETOUR), CANCEL (ANNULER), FINISH (TERMINER) (Le menu de mise en service rapide 1.1 est affich ds la fin de l'assistant (ou lorsque celui-ci est annul). Lorsque CANCEL (ANNULER) est slectionn, aucune valeur n'est enregistre.)

Pour transfrer les valeurs rgles durant la mise en service vers l'appareil et revenir au mode Programmation, utiliser la flche BAS la flche GAUCHE (Terminer) Appuyer trois fois sur pour revenir au mode de Mesure.

N.B. : pour garantir une prcision optimale, il est fortement recommand de raliser un 2.15.2.Auto Zero Head (Hauteur de lame zro auto) la fin de la procdure de mise en service rapide. Pour plus de dtails se reporter la version intgrale des Instructions de service. 1.2. Contrle de pompage
Cet Assistant permet de rgler les pompes utilises dans l'application. Veiller complter d'abord la procdure de mise en service rapide l'aide de l'Assistant appropri.
a) b) c) Appuyer sur la flche DROITE pour lancer le mode PROGRAM et accder au menu niveau 1 : MAIN MENU (MENU PRINCIPAL). Appuyer sur la flche DROITE Appuyer sur la flche BAS pompage. d) A chaque tape, appuyer sur la flche BAS accder directement l'item suivant, ou sur la flche DROITE surligne. e) pour confirmer les valeurs par dfaut et pour accder l'item 1.1 du menu. et sur la flche DROITE pour accder au contrle de

pour accder au mode Edition : la valeur slectionne est pour sauvegarder la modification.

Accder l'item souhait et appuyer sur flche DROITE Appuyer sur flche BAS pour continuer.

f)

A tout moment, vous pouvez appuyer sur la flche HAUT flche GAUCHE pour quitter l'Assistant.

pour revenir en arrire, ou sur la

Assistant de mise en service rapide - Contrle de pompage

Indique le type d'Assistant qui sera lanc.

Options

CANCEL (ANNULER), START (DEMARRER)

Nombre de pompes

Slectionner le nombre de pompes utilis.

Options

NONE (AUCUN(E)), 2 Valeur par dfaut : NONE (AUCUN(E))

Si le rglage est AUCUN(E), le contrle de pompage est dsactiv. 7ML19985XU81 SITRANS LUT400 (HART) Page FR-29

Relais pompe 1

Dfinit le relais attribu la Pompe 1.

Franais

Options
Relais pompe 2

RELAY 2 (RELAIS 2), RELAY 3 Valeur par dfaut : RELAY 2

Visualisation uniquement. Dfinit automatiquement le relais attribu la Pompe 2 en fonction du relais attribu la Pompe 1 dans l'tape prcdente.
Si le relais Pompe 1 = RELAIS 2, alors le relais pompe 2 = RELAIS 3 Options (visualis. Si le relais Pompe 1 = RELAIS 3, alors le relais pompe 2 = RELAIS 2 uniquement)

Mode contrle de pompage

Rgle l'algorithme de contrle associ lactivation du relais.

LUT420 pour le contrle de niveau :


ALTERNATE DUTY ASSIST (POMPAGE CUMULATIF ALTERNE), ALTERNATE DUTY BACKUP (DOUBLE COMMUTATION ALTERNEE)

LUT430 pour le contrle de pompage et de dbit :


ALTERNATE DUTY ASSIST (POMPAGE CUMULATIF ALTERNE), ALTERNATE DUTY BACKUP (DOUBLE COMMUTATION ALTERNEE), SERVICE RATIO DUTY ASSIST (RATIO CUMULATIF ALTERNE), SERVICE RATIO DUTY BACKUP (RATIO FONCT. DOUBLE COMMUTATION), FIXED DUTY ASSIST (POMPAGE CUMULATIF), FIXED DUTY BACKUP (POMPAGE DOUBLE COMMUTATION)

Options supportes par modle

LUT440 pour la mesure de dbit en canal ouvert :


ALTERNATE DUTY ASSIST (POMPAGE CUMULATIF ALTERNE), ALTERNATE DUTY BACKUP (DOUBLE COMMUTATION ALTERNEE), SERVICE RATIO DUTY ASSIST (RATIO CUMULATIF ALTERNE), SERVICE RATIO DUTY BACKUP (RATIO FONCT. DOUBLE COMMUTATION), FIXED DUTY ASSIST (POMPAGE CUMULATIF), FIXED DUTY BACKUP (POMPAGE DOUBLE COMMUTATION)

Valeur par dfaut (tous les modles) : ALTERNATE DUTY ASSIST (CUMULATIF ALTERNE)

Ratio fonctionnement pompe 1

Slectionne les pompes en fonction du nombre dheures de fonctionnement, en non en fonction de la squence normale de pompage.

Valeur

Plage : 0 255 Valeur par dfaut : 1

Ce paramtre est affich uniquement lorsqu'un algorithme de Ratio fonctionnement est slectionn en Mode de contrle de pompage. Ratio fonctionnement pompe 2

Slectionne les pompes en fonction du nombre dheures de fonctionnement, en non en fonction de la squence normale de pompage..

Valeur

Plage : 0 255 Valeur par dfaut : 1

Ce paramtre est affich uniquement lorsqu'un algorithme de Ratio fonctionnement est slectionn en Mode de contrle de pompage.

Page FR-30

SITRANS LUT400 (HART)

7ML19985XU81

Dure de fonctionnement relais 2

Rgler la dure de fonctionnement du relais de pompage 2, en heures.

Valeur

Plage : 0 999999 Valeur par dfaut : 0

Franais

Utiliser la valeur par dfaut pour les pompes intgres posteriori, ou rgler cette valeur pour les pompes existantes prsentant une dure de fonctionnement suffisante. (Ce paramtre est affich uniquement si un algorithme relatif au Ratio de fonctionnement est slectionn en Mode de contrle de pompage.) Dure de fonctionnement relais 3

Rgler la dure de fonctionnement du relais de pompage 3, en heures.

Valeur

Plage : 0 999999 Valeur par dfaut : 0

Utiliser la valeur par dfaut pour les pompes intgres posteriori, ou rgler cette valeur pour les pompes existantes prsentant une dure de fonctionnement suffisante. (Ce paramtre est affich uniquement si un algorithme relatif au Ratio de fonctionnement est slectionn en Mode de contrle de pompage.) Point de consigne pompe ON (1)

Le niveau auquel est actionne la Pompe 1, dfini en 2.1.1.Units (Units).

Valeur

Plage : 0,000 99999,000 Valeur par dfaut : 0.000

Point de consigne pompe ON (2)

Le niveau auquel est actionne la Pompe 2, dfini en 2.1.1.Units (Units).

Valeur

Plage : 0,000 99999,000 Valeur par dfaut : 0.000

Point de consigne pompe OFF (1)

Le niveau auquel est dsactive la Pompe 1, dfini en 2.1.1.Units (Units).

Valeur

Plage : 0,000 99999,000 Valeur par dfaut : 0.000

Point de consigne pompe OFF (2)

Le niveau auquel est dsactive la Pompe 2, dfini en 2.1.1.Units (Units).

Valeur

Plage : 0,000 99999,000 Valeur par dfaut : 0.000

Fin de l'Assistant de mise en service rapide - Contrle de pompage

La bonne excution de la procdure de mise en service suppose la prise en compte des modifications.

Options

BACK (RETOUR) , CANCEL (ANNULER), FINISH (TERMINER) (Le menu de contrle de pompage est affich ds la fin de l'assistant (ou lorsque celui-ci est annul). Lorsque CANCEL (ANNULER) est slectionn, aucune valeur n'est enregistre.)

Pour transfrer les valeurs rgles vers l'appareil et revenir au mode Programmation, utiliser la flche BAS (Terminer). Appuyer trois fois sur la flche GAUCHE pour revenir au mode de Mesure. 7ML19985XU81 SITRANS LUT400 (HART) Page FR-31

Communication SITRANS LUT400 : HART


Pour configurer le systme SITRANS LUT400 via le rseau HART, utiliser soit le programmateur HART Communicator 375/475 Emerson (se reporter la version intgrale des Instructions de service), soit un logiciel ddi. Nous recommandons le logiciel SIMATIC Process Device Manager (PDM) de Siemens. Vous trouverez plus de dtails dans le guide d'application Working with HART, tlchargeable sur la page produit de notre site internet. Consulter : www.siemens.com/sitransLUT400, rubrique Support, cliquez sur Application Guides.

Franais

Commande distance

SITRANS LUT400 est compatible avec diffrents logiciels grce la communication distance : PC quip de SIMATIC PDM PC quip d'AMS Device Manager, de Emerson PC quip d'un navigateur internet PC quip d'un FDT (Field Device Tool) Terminal de Communication 375/475 (FC375/FC475) Les consignes gnrales d'utilisation de ces outils avec le SITRANS LUT400 sont dcrites cidessous. Pour plus de dtails sur les diffrents outils, veuillez consulter les instructions relatives la Communication SITRANS LUT400 (HART).

SIMATIC PDM

Le logiciel SIMATIC PDM est utilis pour la configuration et la maintenance d'appareils tels que le SITRANS LUT400. Pour plus d'informations, voir www.siemens.com/simatic-pdm. Consulter la page produit sur notre site web, pour obtenir la dernire version de SIMATIC PDM, le dernier SP (Service Pack) et le dernier HF (hot fix). Consulter : http://support.automation.siemens.com/WW/ llisapi.dll?func=cslib.csinfo&lang=en&objiD=10806857&subtype=133100

Actualisation de l'Electronic Device Description (EDD)


Vous trouverez la EDD dans la rubrique Device Catalog, sous Sensors/Level/Echo/Siemens AG/ SITRANS LUT400. (La EDD est conue pour garantir la compatibilit ascendante.) Pour faciliter l'identification du fichier EDD, les numros de rvision du fichier EDD et du micrologiciel doivent correspondre (et plus particulirement les numros cls/secondaires, en caractres gras : 1.00.00-02). La rvision EDD doit correspondre la rvision du micrologiciel utilis. Ce numro peut tre vrifi via PDM : pour cela, consulter SITRANS LUT400 HART > Identification > Device.

quivalence rvisions du micrologiciel et du EDD

L'installation de la dernire version de SIMATIC PDM suppose la disponibilit du dernier SP (Service Pack) ainsi que du dernier HF (Hot Fix).

Pour installer un nouveau EDD


Consulter www.siemens.com/sitransLUT400 > Support > Software Downloads pour tlcharger le dernier fichier EDD, disponible sur la page produit de notre site internet. Sauvegarder les fichiers sur votre ordinateur. Extraire le fichier zipp vers un emplacement accessible. Lancer SIMATIC PDM Manager Device Catalog, accder au dossier contenant le fichier dzipp EDD, et le slectionner.

Page FR-32

SITRANS LUT400 (HART)

7ML19985XU81

Configuration d'un nouvel appareil


1) 2) Il est impratif d'utiliser le tout dernier fichier EDD. Vous pouvez, si ncessaire, l'actualiser (cf. Pour installer un nouveau EDD ci-dessus). Configurer le SITRANS LUT400 l'aide de l'Assistant de mise en service adapt votre application.

Franais

Assistants de mise en service via SIMATIC PDM


N.B. :
Les rglages raliss durant la mise en service rapide sont interdpendants. Pour tre prises en compte, les modifications doivent tre valides en cliquant sur Apply and Transfer la fin de la dernire tape. Cliquer sur BACK (RETOUR) pour revenir en arrire et corriger une valeur, ou CANCEL (ANNULER) pour quitter la procdure de mise en service rapide. Accder au menu Device (Appareil) sous SIMATIC PDM, et slectionner l'Assistant de mise en service rapide adapt votre application (Niveau, Volume, Volume-Linarisation, Dbit).

Mise en service rapide

Utilisation avec AMS Device Manager


Le logiciel AMS Device Manager est utilis pour la configuration et la maintenance d'appareils tels que le SITRANS LUT400. Pour plus de dtails : http://www.emersonprocess.com/AMS/.

Electronic Device Description (EDD)


SITRANS LUT400 requiert le fichier EDD pour la version 10.5 du logiciel AMS Device Manager.

Configuration d'un nouvel appareil avec AMS Device Manager


1) Le fichier EDD est disponible dans la rubrique Device Catalog, sous Sensors/Level/Echo/ Siemens AG/SITRANS LUT400. Consulter la page produit sur notre site web : www.siemens.com/sitransLUT400, rubrique Support->Software Downloads, pour vous assurer que vous disposez de la dernire version du fichier EDD pour AMS Device Manager. Sauvegarder les fichiers sur l'ordinateur. Extraire le fichier zipp vers un emplacement accessible. Lancer AMS Device Manager Add Device Type, accder au fichier dzipp EDD et le slectionner. (Un Guide d'application ddi la configuration des appareils HART avec le logiciel AMS Device Manager est disponible sur la page produit de notre site internet, sous Support. Dans le champ Device Connection View (Connexion Appareil), cliquer droite sur l'icne appareil et slectionner Scan Device (vrification appareil) pour tlcharger les paramtres depuis l'appareil. Double-cliquer sur l'icne appareil pour accder l'cran de dmarrage. L'cran de dmarrage indique les donnes d'identification de l'appareil, et une fentre de navigation gauche de l'cran. Configurer le SITRANS LUT400 l'aide de l'Assistant de mise en service adapt votre application. SITRANS LUT400 (HART) Page FR-33

2) 3)

4) 5) 6)

7ML19985XU81

Commande via FC375/475 (HART)


Franais
Le communicateur FC375/FC475 HART est une interface portable trs simple d'utilisation, compatible avec tous les appareils HART tels que le SITRANS LUT400. La structure des menus du FC375/475 est trs similaire celle du logiciel AMS Device Manager.

Electronic Device Description (EDD)


Le SITRANS LUT400 doit tre utilis avec le fichier EDD pour FC375/FC475.

Configuration d'un autre appareil via FC375/475


1) 2) 3) 4) Consulter la page produit de notre site internet, www.siemens.com/sitransLUT400, rubrique Support->Software Downloads, pour disposer de la dernire version du fichier EDD pour FC375/FC475. Sauvegarder les fichiers sur l'ordinateur. Extraire le fichier zipp vers un emplacement accessible. Pour installer le fichier EDD, excuter le logiciel Field Communicator Easy Upgrade. Configurer le SITRANS LUT400 l'aide de l'Assistant de mise en service adapt votre application.

Fonctionnement via FDT (Field Device Tool)


Le standard de fichiers FDT est utilis dans plusieurs logiciels conus pour la configuration et la maintenance d'appareils de terrain comme le SITRANS LUT400. Parmi ces logiciels FDT se trouvent PACTware et Fieldcare. FDT est trs similaire PDM. La configuration d'un appareil de terrain via FDT ncessite le programme DTM (Device Type Manager). Les instruments de mesure Siemens utilisent SITRANS DTM ainsi qu'un fichier de description lectronique EDD pour SITRANS DTM. 1) 2) La premire tape consiste installer SITRANS DTM. Cet application est disponible sur : http://www.siemens.com/sitransdtm. Cliquer sur Support, puis Software downloads. Installer le fichier EDD du SITRANS LUT400 (HART) pour SITRANS DTM. Ce dernier peut tre tlcharg de la page produit de notre site internet : Consulter www.siemens.com/ sitransLUT400 sous Support, et cliquer sur Software Downloads.

Configuration d'un autre appareil via FDT


Un guide d'application est disponible sur la page produit de notre site internet, sous Support.

Entretien et maintenance
Utilis en conditions de fonctionnement normales, le SITRANS LUT400 ne requiert ni maintenance, ni nettoyage.

Rparation de l'instrument et limite de responsabilit


Pour plus de dtails, veuillez vous reporter la dernire page.

Page FR-34

SITRANS LUT400 (HART)

7ML19985XU81

Instructions relatives au cblage pour l'installation en zone dangereuse


Il existe diffrentes options de cblage pour l'installation en zone dangereuse : Cblage en atmosphre explosive (poussire) (International/Europe) Cblage non incendiaire (USA/Canada) Dans tous les cas, se reporter l'tiquette d'identification de l'appareil pour vrifier les homologations.

Franais

1. Cblage en atmosphre explosive (poussire) (Internat./Europe)


SITRANS LUT420 7MLxxxx-xxxxx-xxxx Serial No.: GYZ / B1034567 0518 KCC-REM-S49 Power Rating: 100 230V 15% SITRANSLUT400 50/60 Hz, 36 VA (10 W) Contact Rating: 1A/5A @ 250V , Non-Inductive Operating Temperature: 20C to 50C Enclosure: IP65 / TYPE 4X / NEMA 4X II 3D Ex tc IIIC T70C Dc Ex tc IIIC T70C Dc SIRA 12ATEX9143X IECEx SIR 12.0057X Warning: Possible static hazard. Do not rub or clean on site with a dry cloth. Warning: Do not open when energized or when an explosive dust atmosphere is present. Siemens Milltronics Process Instruments, Peterborough Assembled in Canada with domestic and imported parts SITRANS LUT430 7MLxxxx-xxxxx-xxxx Serial No.: GYZ / B1034567 0518 KCC-REM-S49 Power Rating: 100 230V 15% SITRANSLUT400 50/60 Hz, 36 VA (10 W) Contact Rating: 1A/5A @ 250V , Non-Inductive Operating Temperature: 20C to 50C Enclosure: IP65 / TYPE 4X / NEMA 4X II 3D Ex tc IIIC T70C Dc Ex tc IIIC T70C Dc SIRA 12ATEX9143X IECEx SIR 12.0057X Warning: Possible static hazard. Do not rub or clean on site with a dry cloth. Warning: Do not open when energized or when an explosive dust atmosphere is present. Siemens Milltronics Process Instruments, Peterborough Assembled in Canada with domestic and imported parts SITRANS LUT440 7MLxxxx-xxxxx-xxxx Serial No.: GYZ / B1034567 0518 KCC-REM-S49 Power Rating: 100 230V 15% SITRANSLUT400 50/60 Hz, 36 VA (10 W) Contact Rating: 1A/5A @ 250V , Non-Inductive Operating Temperature: 20C to 50C Enclosure: IP65 / TYPE 4X / NEMA 4X II 3D Ex tc IIIC T70C Dc Ex tc IIIC T70C Dc SIRA 12ATEX9143X IECEx SIR 12.0057X Warning: Possible static hazard. Do not rub or clean on site with a dry cloth. Warning: Do not open when energized or when an explosive dust atmosphere is present. Siemens Milltronics Process Instruments, Peterborough Assembled in Canada with domestic and imported parts

SITRANS LUT420 7MLxxxx-xxxxx-xxxx Serial No.: GYZ / B1034567 0518 KCC-REM-S49 Power Rating: 10 32V , 10W SITRANSLUT400 Contact Rating: 1A/5A @ 250V , Non-Inductive Operating Temperature: 20C to 50C Enclosure: IP65 / TYPE 4X / NEMA 4X

SITRANS LUT430 7MLxxxx-xxxxx-xxxx Serial No.: GYZ / B1034567 0518 KCC-REM-S49 Power Rating: 10 32V , 10W SITRANSLUT400 Contact Rating: 1A/5A @ 250V , Non-Inductive Operating Temperature: 20C to 50C Enclosure: IP65 / TYPE 4X / NEMA 4X

SITRANS LUT440 7MLxxxx-xxxxx-xxxx Serial No.: GYZ / B1034567 0518 KCC-REM-S49 Power Rating: 10 32V , 10W SITRANSLUT400 Contact Rating: 1A/5A @ 250V , Non-Inductive Operating Temperature: 20C to 50C Enclosure: IP65 / TYPE 4X / NEMA 4X

II 3D Ex tc IIIC T70C Dc Ex tc IIIC T70C Dc SIRA 12ATEX9143X IECEx SIR 12.0057X Warning: Possible static hazard. Do not rub or clean on site with a dry cloth. Warning: Do not open when energized or when an explosive dust atmosphere is present.

II 3D Ex tc IIIC T70C Dc Ex tc IIIC T70C Dc SIRA 12ATEX9143X IECEx SIR 12.0057X Warning: Possible static hazard. Do not rub or clean on site with a dry cloth. Warning: Do not open when energized or when an explosive dust atmosphere is present.

II 3D Ex tc IIIC T70C Dc Ex tc IIIC T70C Dc SIRA 12ATEX9143X IECEx SIR 12.0057X Warning: Possible static hazard. Do not rub or clean on site with a dry cloth. Warning: Do not open when energized or when an explosive dust atmosphere is present.

Siemens Milltronics Process Instruments, Peterborough Assembled in Canada with domestic and imported parts

Siemens Milltronics Process Instruments, Peterborough Assembled in Canada with domestic and imported parts

Siemens Milltronics Process Instruments, Peterborough Assembled in Canada with domestic and imported parts

Le certificat ATEX figurant sur l'tiquette d'identification de l'appareil peut tre tlcharg sur la page produit correspondante, sur notre site internet www.siemens.com/sitransLUT400. Le certificat IECEx indiqu sur l'tiquette d'identification de l'appareil est accessible sur le site internet IECEx. Consulter : http://iecex.iec.ch et cliquer sur Ex Equipment Certificates of Conformity. Entrer le numro de certificat : IECEx SIR 12.0057X. Effectuer le cblage selon les dispositions rglementaires locales. Se reporter au certificat ATEX indiqu ci-dessus galement.

7ML19985XU81

SITRANS LUT400 (HART)

Page FR-35

2. Cblage non incendiaire (Canada)


Franais
SITRANS LUT420 7MLxxxx-xxxxx-xxxx Serial No.: GYZ / B1034567 Power Rating: 100 230V 15% 50/60 Hz, 36 VA (10 W) Contact Rating: 1A/5A @ 250V , Non-Inductive Operating Temperature: 20C to 50C Enclosure: IP65 / TYPE 4X / NEMA 4X SITRANS LUT430 7MLxxxx-xxxxx-xxxx Serial No.: GYZ / B1034567 Power Rating: 100 230V 15% 50/60 Hz, 36 VA (10 W) Contact Rating: 1A/5A @ 250V , Non-Inductive Operating Temperature: 20C to 50C Enclosure: IP65 / TYPE 4X / NEMA 4X SITRANS LUT440 7MLxxxx-xxxxx-xxxx Serial No.: GYZ / B1034567 Power Rating: 100 230V 15% 50/60 Hz, 36 VA (10 W) Contact Rating: 1A/5A @ 250V , Non-Inductive Operating Temperature: 20C to 50C Enclosure: IP65 / TYPE 4X / NEMA 4X

159134

Class I, Div.2, Gr. A, B, C & D T3C Class II, Div. 2, Gr. F & G Class III Per Drawing A5E03936871

159134

Class I, Div.2, Gr. A, B, C & D T3C Class II, Div. 2, Gr. F & G Class III Per Drawing A5E03936871

159134

Class I, Div.2, Gr. A, B, C & D T3C Class II, Div. 2, Gr. F & G Class III Per Drawing A5E03936871

Siemens Milltronics Process Instruments, Peterborough Assembled in Canada with domestic and imported parts

Siemens Milltronics Process Instruments, Peterborough Assembled in Canada with domestic and imported parts

Siemens Milltronics Process Instruments, Peterborough Assembled in Canada with domestic and imported parts

SITRANS LUT420 7MLxxxx-xxxxx-xxxx Serial No.: GYZ / B1034567 Power Rating: 10 32V , 10W Contact Rating: 1A/5A @ 250V , Non-Inductive Operating Temperature: 20C to 50C Enclosure: IP65 / TYPE 4X / NEMA 4X

SITRANS LUT430 7MLxxxx-xxxxx-xxxx Serial No.: GYZ / B1034567 Power Rating: 10 32V , 10W Contact Rating: 1A/5A @ 250V , Non-Inductive Operating Temperature: 20C to 50C Enclosure: IP65 / TYPE 4X / NEMA 4X

SITRANS LUT440 7MLxxxx-xxxxx-xxxx Serial No.: GYZ / B1034567 Power Rating: 10 32V , 10W Contact Rating: 1A/5A @ 250V , Non-Inductive Operating Temperature: 20C to 50C Enclosure: IP65 / TYPE 4X / NEMA 4X

159134

Class I, Div.2, Gr. A, B, C & D T3C Class II, Div. 2, Gr. F & G Class III Per DWG. A5E03936871

159134

Class I, Div.2, Gr. A, B, C & D T3C Class II, Div. 2, Gr. F & G Class III Per DWG. A5E03936871

159134

Class I, Div.2, Gr. A, B, C & D T3C Class II, Div. 2, Gr. F & G Class III Per DWG. A5E03936871

Siemens Milltronics Process Instruments, Peterborough Assembled in Canada with domestic and imported parts

Siemens Milltronics Process Instruments, Peterborough Assembled in Canada with domestic and imported parts

Siemens Milltronics Process Instruments, Peterborough Assembled in Canada with domestic and imported parts

Le schma de raccordement CSA Classe I, Div 2 rf. A5E03936871 est disponible sur la page produit de notre site internet : www.siemens.com/sitransLUT400.

Instructions spcifiques pour linstallation en zone dangereuse (rf. Directive Europenne ATEX 94/9/CE, Annexe II, 1/0/6)
Les consignes ci-dessous s'appliquent l'appareil objet du certificat N SIRA 12ATEX9143X: 1) 2) 3) 4) L'utilisation et l'assemblage sont dcrits dans les instructions de service. L'appareil est certifi en tant qu'instrument de la catgorie 3D. L'appareil peut tre utilis en zones atmosphre explosive, zone 22. La temprature la surface de l'appareil ne doit pas dpasser T70 C (temprature ambiante +50C). Slectionner cet appareil en accord avec le code de bonne pratique applicable, suivant des tempratures spcifiques lies l'inflammation de poussires. 5) L'appareil peut tre utilis lorsque la temprature ambiante varie entre -20 et +50 C. 6) Les raccordements des cbles/conduits doivent tre conformes aux conditions tablies par la Directive Europenne 94/9/CE applicable au Groupe II, Catgorie 3D. Ils doivent permettre de garantir la protection IP du botier. 7) Choisir le cble en accord avec les codes de bonne pratique applicables. L'isolation doit notamment pouvoir supporter la temprature de surface maximale spcifie pour le botier (T70 C). 8) Cet appareil n'a pas t valu en tant que dispositif de scurit (conformment la Directive 94/9/CE Annexe II, clause 1.5). 9) L'installation et la vrification de cet appareil doivent tre effectues par un personnel qualifi, en accord avec le code de bonne pratique applicable (EN 60079-14 et EN 6007917 pour l'Europe). 10) Toute rparation de cet appareil doit tre effectue par un personnel qualifi, en accord avec le code de bonne pratique applicable (ex. EN 60079-19 pour lEurope).

Page FR-36

SITRANS LUT400 (HART)

7ML19985XU81

11) Les composants intgrs dans l'appareil ou utiliss pour les remplacements ventuels devront tre installs par un personnel qualifi en accord avec les spcifications contenues dans la documentation fournie par le fabricant. 12) Pour dclencher une coupure manuellement, utiliser le coupe-circuit prvu dans l'installation.

Franais

CONDITIONS SPECIALES POUR UNE UTILISATION SRE


Le suffixe X du numro de certificat dsigne les conditions spciales suivantes, permettant lutilisation de l'appareil en toute scurit : Les raccordements de cbles/conduits doivent tre conformes aux conditions de la Directive Europenne 94/9/CE applicable aux appareils du Groupe II, Catgorie 3D, et permettre de garantir la protection IP du botier. Choisir le cble en accord avec les codes de bonne pratique applicables. L'isolation doit pouvoir supporter la temprature de surface maximale spcifie pour le botier (T70 C). Le botier non-conducteur peut, sous certaines conditions extrmes, tre l'origine de charges lectrostatiques suffisantes pour dclencher l'inflammation. L'emplacement choisi pour l'installation doit permettre de protger l'instrument des conditions externes (vapeur haute pression par exemple) qui peuvent donner lieu une accumulation de charge lectrostatique sur les surfaces non-conductrices. Par ailleurs, l'instrument doit tre nettoy avec un chiffon humide uniquement. L'instrument doit tre install l o le risque d'endommagement mcanique est moindre ou, alternativement, protg par un dispositif adapt, permettant de garantir la protection mcanique de ce dernier. Les contrleurs de la srie LUT400 (LUT420, LUT430, LUT440) utiliss avec un couvercle pour montage panneau dport doivent tre installs dans un botier de protection adapt, rpondant aux conditions de la Directive Europenne 94/9/CE applicable aux appareils du Groupe II, Catgorie 3D, IP65 (minimum).

7ML19985XU81

SITRANS LUT400 (HART)

Page FR-37

Notes
Franais
Page FR-38

SITRANS LUT400 (HART)

7ML19985XU81

Manuale per l'avvio rapido del SITRANS LUT400 (HART)


Questo manuale descrive le caratteristiche e le funzioni pi importanti del SITRANS LUT400 (HART 1)). Si consiglia all'operatore di leggere il manuale d'uso integrale per ottenere i massimi risultati. La versione intagrale delle Istruzioni d'uso SITRANS LUT400 (HART) e il manuale per la Comunicazione SITRANS LUT400 (HART) sono scaricabili dal sito: www.siemens.com/ sitransLUT400. Per la versione stampata, rivolgersi al proprio rappresentante di Siemens. Per ulteriori informazioni su questo manuale, rivolgersi a: Siemens AG Siemens Milltronics Process Instruments 1954 Technology Drive, P .O. Box 4225 Peterborough, Ontario, Canada, K9J 7B1 Email: techpubs.smpi@siemens.com

Italiano

Copyright Siemens AG 2012. Tutti i diritti riservati


Si consiglia agli utenti di acquisire manuali stampati autorizzati o di consultare le versioni elettroniche progettate e realizzate da Siemens Milltronics Process Instruments. Siemens Milltronics Process Instruments non potr essere ritenuta responsabile per la riproduzione parziale o totale del contenuto delle versioni stampate o elettroniche.

Clausola di esclusione della responsabilit


La conformit tra lo stato tecnico dell'apparecchio e il contenuto di questo manuale stata verificata; tuttavia, potrebbero essere riscontrate alcune variazioni. SMPI non garantisce pertanto la completa conformit del manuale con l'apparecchio descritto. Si avvisa inoltre che tutti i manuali vengono regolarmente controllati ed aggiornati e che le eventuali correzioni vengono incluse nelle versioni successive. Si invitano gli utenti a trasmettere i propri suggerimenti e commenti a SMPI. SMPI si riserva il diritto di apportare modifiche tecniche.

MILLTRONICS un marchio registrato di Siemens Milltronics Process Instruments.

Supporto tecnico
Il servizio disponible 24 ore su 24. Per trovare l'ufficio di rappresentanza Siemens Automation pi vicino, consultare il sito web: www.siemens.com/automation/partner: Cliccare sul tab Contact per scegliere il prodotto (ad esempio: Tecnica di automazione). Cliccare nuovamente sull'elenco a discesa dei prodotti per consultare la famiglia di prodotti. Per la misura di livello: +Tecnica di automazione > +Sistemi di sensori >+Strumentazione di processo > +Misura di livello > +Misura di livello continua. Selezionare A ultrasuoni. Per la misura di portata: +Tecnica di automazione > +Sistemi di sensori >+Strumentazione di processo > +Misura di portata. Selezionare A ultrasuoni. Selezionare un paese e una citt/Regione. Selezionare Technical Support nell'elenco Servizi. Per accedere al supporto tecnico on-line, visitare: www.siemens.com/automation/support-request Introdurre il nome dell'apparecchio (SITRANS LUT400) o il numero dell'ordine di acquisto, cliccare su Search (Cerca), e selezionare il tipo di prodotto. Cliccare su Next. Introdurre un keyword per la richiesta. Visualizzare la documentazione o cliccare suNext (Avanti) per inviare una descrizione della richiesta via e-mail al nostro team di Supporto Tecnico. Centro di Supporto Tecnico Siemens IA/DT: tel. +49 (0)911 895 7222
1)

HART un marchio registrato di HART Communication Foundation.

7ML19985XU81

SITRANS LUT400 (HART)

Pagina IT-1

Indicazioni di sicurezza
L'apparecchio deve essere utilizzato osservando le avvertenze. Un utilizzo non corretto potrebbe causare danni anche gravi alle persone, sia al prodotto e agli apparecchi ad esso collegati. Le avvertenze contengono una spiegazione dettagliata del livello di sicurezza da osservare. Il simbolo AVVERTENZA sul prodotto indica all'utente che la mancata osservazione delle precauzioni necessarie pu causare morte o gravi lesioni personali nonch seri danni materiali. Il simbolo AVVERTENZA usato quando sul prodotto non riportato un simbolo di avvertenza. Indica all'utente che la mancata osservazione delle precauzioni necessarie pu causare morte o gravi lesioni personali nonch seri danni materiali.

Italiano

Nota: le note contengono importanti informazioni sul prodotto o sulla sezione del manuale d'istruzioni a cui viene fatto riferimento.

Conformit FCC
Installazioni negli USA: Regole della FCC (Federal Communications Commission)
AVVERTENZA: Ogni cambiamento o modifica non espressamente approvato da Siemens Milltronics potrebbe revocare il diritto all'utilizzo dell'apparecchio. Note: Il presente apparecchio stato collaudato e risultato conforme ai vincoli relativi ai dispositivi digitali di Classe A, inclusi nella Parte 15 della normativa FCC. Tali vincoli sono stati stabiliti ai fini di garantire un'adatta protezione da interferenze dannose durante l'utilizzo dell'apparecchio in ambienti commerciali. Il presente apparecchio genera, utilizza e pu emettere energia in radio frequenza. Se installato e utilizzato in modo improprio, vale a dire in difformit con il relativo manuale d'istruzioni, il presente apparecchio pu causare interferenze dannose alle radiocomunicazioni. L'impiego del presente apparecchio all'interno di una zona residenziale potrebbe causare interferenze dannose. In tal caso, l'utente dovr risolvere il problema ed eliminare tali interferenze a proprie spese.

Pagina IT-2

SITRANS LUT400 (HART)

7ML19985XU81

SITRANS LUT400
AVVERTENZA: Per garantire la protezione il SITRANS LUT400 deve essere utilizzato osservando le istruzioni e avvertenze contenute in questo manuale. Nota: Questo prodotto stato progettato per l'uso in ambienti industriali. Luso di questo apparecchio in zone residenziali pu causare disturbi alle comunicazioni radiofoniche. I controllori di livello ad ultrasuoni SITRANS LUT400 della Siemens sono apparecchi compatti per misure singole di lunga portata. Sono adatti per la misura di liquidi, solidi o fanghi e la misura di portata ad alta precisione su canali aperti. La serie SITRANS LUT400 compatibile con l'intera gamma dei trasduttori EchoMax della Siemens, con un range di misura da 0,3 a 60 metri (a seconda del trasduttore). SITRANS LUT400 dispone di un'interfaccia utente con display retroilluminato (LUI) per la programmazione tramite men. E' inoltre dotato di un'interfaccia web intuitiva e di assistenti personalizzati per consentire l'esecuzione plug & play. SITRANS LUT400 compatibile con il protocollo di comunicazione HART, e diversi software per la comunicazione remota [SIMATIC PDM, AMS Device Manager, Field Communicator 375/475, e FDT (Field Device Tools)]. E' dotato della tecnologia di elaborazione dell'eco Sonic Intelligence, che garantisce prestazioni eccezionali e facilit d'uso.

Italiano

Specifiche tecniche
Alimentazione

Il dati tecnici completi sono descritti nelle Instruzioni operative SITRANS LUT400 (HART).

Modello in corrente alternata (AC)

100-230 V AC 15%, 50 / 60 Hz, 36 VA (10W)1) Fusibile: 5 x 20 mm, lento, 0.25A, 250V


E indicata la massima corrente assorbita.

Modello in corrente continua (DC)


10-32 V DC, 10W 1 Fusibile: 5 x 20 mm, lento, 1,6A, 125V

1)

Condizioni ambientali
Altitudine
2000 m max

Temperatura ambiente
da -20 a +50 C (-4 a +122 F)

Categoria di installazione
II

Grado di inquinamento
4

Umidit relativa
Adatto per l'installazione in esterni (solo con custodia IP65 Tipo 4X / NEMA 4X)

Certificazioni
Note: Le omologazioni sono indicate sull'etichetta dell'apparecchio. Utilizzare guarnizioni adatte a tenuta stagna per garantire il grado di protezione IP o NEMA. Uso generale Aree a rischio esplosione Non-incendive (Canada) A prova di esplosione (polvere) (Internazionale/Europa) CSAUS/C, CE, FM, UL listed, C-TICK CSA Classe I, Div. 2, Gr. A, B, C, D; Classe II, Div 2, Gruppi F, G; Classe III ATEX II 3D Ex tc IIIC T70 C IP65 IECEx SIR 12.0057X SIRA 12ATEX9143X

7ML19985XU81

SITRANS LUT400 (HART)

Pagina IT-3

Installazione
Note: L'installazione deve essere eseguita esclusivamente da personale qualificato e in conformit con le normative locali correnti. Questo apparecchio sensibile alle scariche elettrostatiche. Garantire una corretta messa a terra. Tutti i collegamenti elettrici devono disporre di adeguato isolamento (250 V min.). Durante il funzionamento i morsetti dei trasduttori sono sottoposti ad una tensione pericolosa. I terminali di ingresso CC devono essere alimentati da una fonte dotata di isolamento elettrico tra l'ingresso e l'uscita, per garantire la conformit ai requisiti di sicurezza applicabili della direttiva IEC 61010-1. I morsetti di collegamento dei rel sono previsti per l'uso con apparecchiature senza parti sotto tensione accessibili, e collegamenti elettrici che dispongono di adeguato isolamento (250 V min). La massima tensione di funzionamento applicabile tra i contatti dei rel 250 V. La custodia non metallica non fornisce un collegamento a terra continuo tra i conduttori. Utilizzare pertanto boccole e ponticelli di messa a terra.

Italiano

Istruzioni generali di montaggio


Le istruzioni di montaggio variano in base alla configurazione scelta: montaggio a muro, su palina, tramite guida DIN o a pannello display remoto. Seguire le istruzioni specifiche per il particolare modello.

Avvertenza: Prima del montaggio del SITRANS LUT400, necessario seguire le istruzioni di Disposizione dei cavi pagina 9. Esecuzione a muro o pannello
Tutti i modelli SITRANS LUT400 sono forniti con una piastra di montaggio posteriore. Il modello per montaggio a pannello viene fornito con coperchio con display LUI e con un coperchio cieco.
piastra di montaggio posteriore

custodia con coperchio dotato di display opzionale

clip di fissaggio (x 2) clip di fissaggio (x 2)

Nota: Elementi di fissaggio per montaggio a muro non forniti.

Pagina IT-4

SITRANS LUT400 (HART)

7ML19985XU81

piastra di montaggio posteriore

custodia con coperchio cieco

clip di fissaggio (x 2) clip di fissaggio (x 2)

Italiano

Nota: Elementi di fissaggio per montaggio a muro non forniti. retro della custodia

attacchi fessurati

Montaggio a parete della custodia


1) 2) 3) Segnare e perforare quattro fori nella superficie di montaggio (fornita dal cliente). Fissare con l'aiuto di un cacciavite. Allineare gli attachi fessurati sulla parte posteriore dell'apparecchio con i clip situati sulla piastra di montaggio posteriore. Spingere il LUT400 contro la piastra posteriore, facendolo scorrere verso il basso per posizionarlo correttamente.

fori per viti di montaggio sulla piastra posteriore

Nota:
Dimensioni consigliate per gli elementi di fissaggio: vite M8 o 5/16" con rondella diametro esterno 17 mm o 5/8" max. Montaggio consigliato: fissaggio diretto a parete. Se l'apparecchio viene fissato su un'altra superficie questa DOVRA' sopportare un peso equivalente a quattro volte quello dell'apparecchio.

7ML19985XU81

SITRANS LUT400 (HART)

Pagina IT-5

Coperchio per montaggio remoto


coperchio display per montaggio remoto

guarnizione

Italiano

4 elementi di fissaggio pannello (forniti)

cavo di prolunga SITRANS LUT400

Installazione del coperchio per montaggio remoto


Nota: Il coperchio per montaggio remoto pu essere installato a una distanza max. di 5 m dall'apparecchio tramite due cavi di prolunga opzionali (ognuno lungo 2,5 m). Consultare la versione integrale delle Istruzioni operative. 1) Applicare il modello fornito e ritagliare il coperchio remoto con display LUI per ottenere l'apposito foro. Collocare la guarnizione all'interno del coperchio, allineando i fori di montaggio. Allineare la parte posteriore del coperchio dotato di display con l'apertura del pannello. Segnare e perforare quattro fori nella superficie di montaggio (fornita). Fissare con l'aiuto di un cacciavite. Nota: Torsione consigliata per l'apposita guarnizione delle viti di fissaggio : 1,1 N m 10 in-lbs

2)

Nota:
Montaggio consigliato: montaggio a pannello, distanza max. di 5 m dall'apparecchio se l'apparecchio viene fissato su un'altra superficie questa DOVRA' sopportare un peso equivalente a quattro volte quello dell'apparecchio. Nota: Elementi di fissaggio forniti: vite M5, rondella di tenuta stagna, rondella piatta e vite M5. Questi elementi di fissaggio sono necessari per garantire il grado di protezione IP65 del coperchio per montaggio remoto. Pagina IT-6 SITRANS LUT400 (HART) 7ML19985XU81

Montaggio su palina

piastra di montaggio posteriore attacco a morsetto custodia con coperchio dotato di display opzionale

bullone aU

Italiano

tubo Nota: Elementi di fissaggio per montaggio su palina non forniti.

Montaggio della custodia


1) 2) 3) Fissare la piastra di montaggio posteriore al tubo tramite bulloni a U e attacchi a morsetto (forniti dal cliente) adatti al diametro del tubo. Fissare i bulloni con una chiave adatta. Il serraggio eccessivo pu piegare e danneggiare la piastra. Questo pu impedire il montaggio del LUT400 sulla piastra posteriore. Fissare l'apparecchio sulla piastra di montaggio posteriore (come descritto nella tappa 3 di Montaggio a parete della custodia pagina 5).

Nota:
Montaggio consigliato: direttamente sul tubo orizzontale o verticale. Se l'apparecchio viene fissato su un'altra superficie questa DOVRA' sopportare un peso equivalente a quattro volte quello dell'apparecchio. Dimensioni consigliate del tubo: max.: tubo 3", min.: tubo 3/4" Dimensioni consigliate per gli elementi di fissaggio: Bulloni a U: max.: tubo 3" con filettatura M8 o 3/8" min.: tubo 3/4" con filettatura M6 o 1/4" Viti a testa esagonale: da M6 o 1/4" a M8 o 3/8" Rondella: max.: diametro esterno 16 mm o 13/16".

7ML19985XU81

SITRANS LUT400 (HART)

Pagina IT-7

Montaggio tramite guida DIN


retro della custodia

Guida DIN

clip di fissaggio (x 2)

fessura

Slitta di scorrimento guida DIN

Italiano Montaggio della custodia


1) 2) 3) Inclinare la parte superiore della custodia verso la guida DIN, posizionandola leggermente sopra la parte superiore della guida. Spostare la custodia verso il basso contro la guida DIN, agganciando i clip situati sulla parte posteriore della custodia alla parte superiore della guida DIN. Spingere l'apparecchio contro la guida DIN, per innestarlo sulla slitta di scorrimento della guida DIN. Questo consente un fissaggio sicuro della custodia sulla guida DIN.

Nota:
Montaggio consigliato: diretto su guida DIN orizzontale. Dimensioni richieste per la guida DIN: TH 35-7.5 o TH 35-15 a norma IEC 60715. La guida DIN DOVRA' sopportare un peso equivalente a quattro volte quello del SITRANS LUT400.

Smontaggio della custodia


1) Partendo dalla parte anteriore dell'apparecchio, introdurre il cacciavite nella fessura situata sulla parte inferiore della slitta di scorrimento della guida DIN. Questo consente di sganciare la slitta di scorrimento dalla parte inferiore della guida DIN. Tenere abbassata la slitta di scorrimento e spingere verso l'alto la custodia per disinnestare le clip dalla parte superiore della guida DIN.

2)

Pagina IT-8

SITRANS LUT400 (HART)

7ML19985XU81

Preparazione all'installazione dei cavi


I cavi possono essere inseriti in conduit di protezione o direttamente all'interno della custodia con pressacavi adatti. Preparare i fori di ingresso dei cavi seguendo le istruzioni sottostanti (tappe 1 - 5). Seguire poi le indicazioni relative all'uso di conduit o di pressacavi. 1) Assicurarsi che il coperchio della custodia sia chiuso e che le viti di fissaggio siano bloccate. 2) Inserire la punta del cacciavite nella scanalatura sulla parte esterna della giunzione preforata (vedi illustrazione sottostante). 3) Colpire l'estremit del cacciavite con il palmo della mano per rimuovere la giunzione. 4) Allentare le viti e rimuovere il coperchio della custodia. 5) Rimuovere i(l) componenti(e) di plastica usati per coprire gli ingressi cavi presenti sulla custodia. Agire con cautela per evitare di danneggiare l'elettronica con l'elettricit statica o con gli utensili impiegati per rimuovere le guarnizioni.
SITRANS LUT400

ingressi cavi

guarnizione preforata (x 3)

Note: Il diametro del foro d'ingresso ottenuto rimuovendo le guarnizioni preforate varia da 21,4 mm a 21,6 mm. I pressacavi M20 (diametro 20 mm) e conduit 1/2" NPT (diametro 21,3 mm) si adattano perfettamente a questo ingresso cavi. Scegliere con cura la guarnizione adatta per ogni ingresso cavo. Si consiglia l'uso di guarniscanalatura zioni piatte (e non di anelli di per cacciavite tenuta). L'uso di elementi di fis(x 3) saggio diversi rende l'utente unico e diretto responsabile di garantire il grado di protezione IP54 degli ingressi cavi.

Italiano

Cavo posizionato all'interno del conduit/tubo


(segue da sopra, tappe 1 - 5) 6) Una volta ultimata la preparazione dell'installazione dei cavi seguendo la indicazioni sopra riportate, fissare il conduit alla custodia con raccordi a tenuta stagna di dimensioni adatte. (Dimensione conduit: 1/2" NPT). 7) Chiudere il coperchio della custodia e stringere le viti.

Cavo esposto, inserito nel pressacavo


(segue da sopra, tappe 1 - 5) 6) Una volta ultimata la preparazione dell'installazione dei cavi seguendo le indicazioni sopra riportate, svitare i pressacavi e fissarli saldamente alla custodia. 7) Inserire i cavi nei pressacavi. Mantenere separati il cavo dell'alimentazione e i cavi di segnale. Collegare i cavi ai blocchi terminali. 8) Stringere i pressacavi per garantire la tenuta stagna. 9) Chiudere il coperchio della custodia e stringere le viti. Pagina IT-9

7ML19985XU81

SITRANS LUT400 (HART)

Collegamenti elettrici
Precauzioni di sicurezza per il cablaggio
I collegamenti elettrici sono descritti in dettaglio nel manuale d'uso integrale. Loperatore deve installare tutti i componenti dellapparecchio secondo le istruzioni. Collegare le schermature dei cavi ai morsetti per schermature del LUT400 (indicati con il ). Per evitare differenze di potenziale tra punti di messa a terra gli schermi simbolo dei cavi vanno opportunamente collegati a terra. I conduttori esposti dei cavi blindati devono essere il pi corto possibile, per limitare le interferenze elettriche causate da trasmissioni o rumore.

Collegamenti elettrici SITRANS LUT400


Italiano
AVVERTENZE:

Per informazioni sulle omologazioni, consultare l'tichetta dell'apparecchio. Utilizzare guarnizioni adatte a tenuta stagna per garantire il grado di protezione IP o NEMA.
Note: Potr essere necessario separare i cavi e conduit in conformit con le procedure standard di cablaggio della strumentazione o con codici locali per materiali elettrici. Per accedere al vano di cablaggio: 1) 2) 3) Svitare di 1/4 di giro le viti di bloccaggio. Sollevare il coperchio a cerniera e aprirlo verso sinistra. Il coperchio pu essere lasciato aperto, sostenuto dalla cerniera, o pu essere sganciato e lasciato staccato per accedere al vano di cablaggio.

4) 5) 6)

Realizzare i collegamenti elettrici seguendo attentamente le indicazioni sottostanti. Una volta terminato il cablaggio, rimettere a posto il coperchio dell'apparecchio. Stringere le viti di bloccaggio.

Pagina IT-10

SITRANS LUT400 (HART)

7ML19985XU81

Vano di cablaggio
La morsettiera del SITRANS LUT400 offre la possibilit di collegare simultaneamente tutti gli ingressi e le uscite. E' inoltre possibile rimuovere i blocchi terminali per facilitare ulteriormente il cablaggio. AVVERTENZA: Accertarsi che i blocchi terminali siano collegati correttamente durante la reinstallazione. La mancata osservanza delle istruzioni pu causare danni all'apparecchio e a sistemi esterni ad esso collegati. Nota: Coppia di serraggio consigliata (viti di serraggio terminali): 0,56 - 0,79 N m 5 - 7 in-lbs Non serrare le viti a fondo. alimentazione
TS-3 TRANSDUCER DISCRETE INPUTS INPUT POWER

Italiano

alimentazione

8 1

9 2

10

11 0.25 A, 250 V SLOW BLOW

12 L1

13 L2/N 100 - 230 V

14

DC SYNC

SHLD

WHT

BLK

BLK

WHT

blocchi terminali
mA OUT/HART BATTERY BR2032

batteria

RELAYS

DISPLAY 15 16 RLY 1 17 18 19 20 21 22 23 24

RLY 2

RLY 3 ACT

PAS SERVICE

connessione USB

cavo display

connessione mA HART

Alimentazione
AVVERTENZE: I terminali di ingresso CC devono essere alimentati da una fonte dotata di isolamento elettrico tra l'ingresso e l'uscita, per garantire la conformit ai requisiti di sicurezza applicabili della direttiva IEC 61010-1. Tutti i collegamenti elettrici di campo devono disporre di adeguato isolamento per le tensioni di esercizio. Nota ! Prima di collegare la tensione di alimentazione del SITRANS LUT400 per la prima volta, assicurarsi che i sistemi di allarme/controllo rimangano disattivati fino ad ottenere risultati soddisfacenti.

7ML19985XU81

SITRANS LUT400 (HART)

Pagina IT-11

Il SITRANS LUT400 disponibile in versione AC e DC.

Alimentazione AC
INPUT POWER

Alimentazione DC
INPUT POWER

12

13

14

12 10 - 32 V

13

14

L1 L2/N 100 - 230 V

AC: 100-230 V AC 15%, 50/60 Hz, 36 VA (10W)

DC: 10-32 V DC, 10W

Italiano

Nota: Eseguire unadeguata messa a terra dellapparecchio. 1) 2) 3) Rimuovere il rivestimento del cavo per circa 70 mm (2.75") partendo dall'estremit, quindi passare il fili attraverso il pressacavo 1). Collegare i fili con gli appositi morsetti seguendo il diagramma sottoriportato: la polarit viene indicata sul blocco terminale. Effettuare il collegamento a terra dell'apparecchio attenendosi alle prescrizioni vigenti. Indicazioni per il collegamento con corrente CA: Proteggere lapparecchio con un fusibile 15 A o un interruttore limitatore di corrente. Accertarsi che nell'installazione e in una posizione facilmente accessibile dall'operatore si trovi un interruttore limitatore di corrente, contrassegnato come sezionatore.

Batteria dell'apparecchio
Il SITRANS LUT400 viene fornito con una batteria. (La posizione della batteria illustrata a pagina 11.) La batteria BR2032 ha una lunga durata di vita (circa 10 anni) ed sensibile alla temperatura ambiente. In caso di interruzione dell'alimentazione elettrica del LUT400, l'orologio in tempo reale alimentato dalla batteria fino al ripristino dell'alimentazione elettrica. (Per sostituire la batteria, consultare la versione integrale delle Istruzioni operative.)

1)Se il cavo viene inserito in un tubo protettivo o conduit, utilizzare un pressatubo di dimensioni adatte per

garantire la tenuta stagna.

Pagina IT-12

SITRANS LUT400 (HART)

7ML19985XU81

Trasduttori
Avvertenza: Durante il funzionamento i morsetti dei trasduttori sono sottoposti ad una tensione pericolosa. Sistemare il cavo del transduttore separatamente (eccetto il cavo del sensore di temperatura TS-3), all'interno di tubi protettivi di metallo con messa a terra. Note: Non utilizzare un cavo coassiale con il SITRANS LUT400 per via dell'alta tensione trasmessa dallo schermo del cavo. Lo schermo del LUT400 e il cavo bianco del transduttore devono essere collegati separatamente utilizzando appositi morsetti. Le istruzioni contenute nelle versioni precedenti delle Istruzioni operative (trasduttori) sono contraddittorie e non vanno seguite.

Italiano

TS-3

TRANSDUCER

SHLD

BLK

WHT

Rel
Gli schemi sono rappresentati con lo stato dei contatti rel diseccitati. Tutti i rel sono configurabili in logica positiva o negativa (per ulteriori informazioni vedi Parametri 2.8.11.Relay Logic nella versione integrale delle Istruzioni Operative).
RELAYS

BLK

WHT

Interruzione dell'alimentazione elettrica


20 21

15

16 RLY 1

17

18

19

RLY 2

RLY 3

Portata contatti rel


un rel Form C (NA o NC) (rel 1), 1A/250 V AC (carico non induttivo) 3A/30 V DC due rel Form A (NA) (rel 2, 3), 5A/250 V AC (carico non induttivo) 3A/30 V DC

Rel 2, 3 normalmente aperti. Lo stato del contatto rel 1 pu essere definito normalmente aperto o normalmente chiuso. Nel caso che si verifichi un'interruzione dell'alimentazione elettrica, i rel tornano al loro stato normale.

7ML19985XU81

SITRANS LUT400 (HART)

Pagina IT-13

Collegamenti HART Uscita mA (HART)


mA OUT/HART

Per un collegamento HART ATTIVO (con fonte di alimentazione elettrica LUT400), collegare i morsetti 22 e 23.

22

23

24

ACT

PAS

Per un collegamento HART PASSIVO (con fonte di alimentazione elettrica esterna), collegare i morsetti 23 e 24.

Per ulteriori dettagli, consultare i parametri di uscita analogica (2.5.Current Output) riportati nella versione integrale delle Istruzioni operative, sezione Parametri.

Italiano

Cablaggio in aree potenzialmente esplosive


Vedi pagina 35.

Messa in servizio del SITRANS LUT400


Consigliamo di eseguire la configurazione dell'apparecchio nell'ordine indicato: Per prima cosa, avviare l'Assistente per la vostra applicazione (Livello, Volume, Portata). Poi, configurare le pompe con l'aiuto dell'Assistente Controllo Pompe (se applicabile). Gli allarmi o altri controlli, totalizzatori e campionatori vanno configurati per ultimo, specificando i relativi parametri (vedi versione integrale delle istruzioni operative). Esistono due modi per accedere agli assistenti: accesso locale (vedi Assistenti di avvio rapido con interfaccia LUI pagina 17) accesso da una postazione remota (vedi Altri Assistenti di avvio rapido: pagina 17)

Messa in servizio del SITRANS LUT400


Collegare l'alimentazione. SITRANS LUT400 si avvia automaticamente nel modo di Misura. Durante l'elaborazione del primo segnale di misura, sul display LUI appare il logo Siemens seguito dall'attuale versione del firmware. Durante la prima configurazione possibile selezionare la lingua (Inglese, Tedesco, Francese, Spagnolo, Cinese). Prima di proseguire con la configurazione dell'apparecchio ' importante impostare la data et l'ora. (Per modificare questi valori sufficiente programmare il parametro 2.14.Date and Time. Consultare la versione integrale delle Istruzioni operative.)

Pagina IT-14

SITRANS LUT400 (HART)

7ML19985XU81

Display LCD Visualizzazione Modo di misura: Funzionamento normale 1) Livello


1 10
SITRANS LUT400 LEVEL

2
1/2

4 5

18.91
1

6
2

5 SAVING PARAMETERS

Portata
1 10 11 12
SITRANS LUT400 FLOW HEAD
1 2 3

3
1/2
L/S M

4 5 6

500313..
60.000

DAILY TOTAL,...

999999.99

5 SAVING PARAMETERS

1 tag 2 valore misurato (livello, spazio, distanza, volume, portata, o battente) 3 valore visualizzato [Variabile Primaria (VP)=1 di 2, Variabile Secondaria (VS)=2 di 2] 4 indicatore di commutazione1) per VP o VS 5 unit 6 diagramma a barre indica il livello 7 la sezione secondaria del display indica i rel configurati (sinistra) e gli ingressi digitali (destra) 8 l'area di visualizzazione testo indica i messaggi di stato 9 indicatore di stato dell'apparecchio 10 modo sensore selezionato (primario): livello, spazio, distanza, volume, battente o portata 11 modo sensore secondario = battente, se il sensore funziona nel modo "portata" 12 valori totalizzatore: display alterna visualizzazione tra totalizzatore giornaliero e totalizzatore operativo

Italiano

Presenza guasto
0 LOE

8 zona di testo, indica un codice di errore sotto forma di un messaggio 9 appare l'icona per indicare che l'apparecchio necessita di manutenzione
men attuale/ nome del parametro

Visualizzazione in modo Program Visualizzazione durante la navigazione

TRANSDUCER 2.1.6 di para Se la barra dei men visibile, la lista NO TRANSDUCER metro XRS-5 dei men troppo lunga e non si XPS-10 visualizzano tutti gli item. XPS-15 banda item XCT-8 Lo larghezza della banda di item nella di item attuale XCT-12 barra dei men riflette la lunghezza XPS-30 barra dei della lista dei men: l'elenco di item men sar tanto pi breve quanto maggiore sar la barra. La posizione della banda di item riflette la posizione approssimativa dell'item selezionato. Se la banda situata nel centro della barra dei men, l'item si trova a met della lista. L'elenco di item sar tanto pi breve quanto maggiore sar la dimensione della barra.

numero

1)Premere

per alternare il display.

7ML19985XU81

SITRANS LUT400 (HART)

Pagina IT-15

Visualizzazione parametri
men del parametro
SENSOR
UNITS SENSOR MODE SENSOR MODE SECONDARY

Modo editazione
2.1.6

numero di parametro valore/ selezione del parametro

TRANSDUCER
NO TRANSDUCER XPS-10 XPS-15 XCT-8 XCT-12 XPS-30

2.1.6

XRS-5

TRANSDUCER

nome del parametro

FREQUENCY

XRS-5

Modo PROGRAM
Con i tasti dell'apparecchio, premere livello 1. Visualizzare il men con , per passare al modo Program e accedere al men , , .

Selezione di un'opzione Italiano


a) Accedere al parametro desiderato. b) Premere nuovamente la freccia DESTRA selezione attuale evidenziata. c) Accedere a una selezione nuova. d) Premere la freccia DESTRA per confermare. La nuova selezione viene visualizzata sull'indicatore LCD. per accedere al modo Editazione . La

Modificazione di un valore numerico


a) Accedere al parametro desiderato, ad esempio Punto di calibrazione basso (2.2.1.) b) Se selezionato, consente di visualizzare il valore attuale. c) Premere la freccia DESTRA per attivare il modo Editazione. La posizione attuale del cursore evidenziata. d) Premere le frecce SINISTRA e DESTRA per posizionare il cursore sulla cifra che si vuole impostare. e) Per ogni cifra selezionata, utilizzare le frecce ALTO o ridurre il suo valore. f) e BASSO per aumentare e BASSO
LOW CALIB. PT. 2.2.1

ESC

+ _ _ _ _ _ _ 2. 0 M

OK

Quando la virgola decimale selezionata, utilizzare le frecce ALTO per spostare la virgola. la freccia SINISTRA

g) Per uscire dal programma senza memorizzare le modifiche, premere continuamente fino a selezionare ESC Premere di nuovo la freccia SINISTRA per uscire dal programma senza memorizzare le modifiche. Altrimenti, se il nuovo valore del parametro corretto, premere continuamente la freccia DESTRA fino a selezionare OK.

h) Premere la freccia DESTRA per confermare. La nuova selezione viene visualizzata sull'indicatore LCD. Verificare che i valori siano conformi.

Pagina IT-16

SITRANS LUT400 (HART)

7ML19985XU81

Assistenti di avvio rapido


Per configurare il SITRANS LUT400 per la misura di livello, volume (forme di serbatoio standard), o portata, utilizzare il Assistenti di avvio rapido con interfaccia LUI descritto sotto. Un'assistente dedicato alle applicazioni con serbatoi di forme pi complesse disponibile con SIMATIC PDM. Vedi Accesso agli assistenti di avvio rapido tramite SIMATIC PDM pagina 33.

Altri Assistenti di avvio rapido:


Sono inoltre disponibili altri assistenti di avvio rapido compatibili con diversi software 1): SIMATIC PDM (HART) AMS (HART) FC375/475 (HART) FDT (Field Device Tool) (HART) Le tabelle di configurazione parametri (per registrare i dati prima di configurare l'apparecchi con un'assistente di avvio rapido) sono disponibili sul nostro sito web. Vedi www.siemens.com/sitransLUT400 > Support (Supporto)> Application Guides (Guide all'uso).

Italiano

Assistenti di avvio rapido con interfaccia LUI


1) 2) Premere per passare al modo Program. Scegliere Assistenti (1.), Avvio rapido (1.1.), e quindi l'Assistente di avvio rapido: Avvio rapido - Livello (1.1.1.), Avvio rapido - Volume (1.1.2.), o Avvio rapido - Portata (1.1.3.). [L'Assistente di avvio Rapido Portata viene indicato sull'interfaccia LUI per il LUT430 (Controllo pompe e Portata), e per il LUT440 (OCM; misura di portata in canale aperto); solo modelli impostati.] Seguire le tappe indicate e selezionare Finish (Fine) per memorizzare i valori impostati e accedere nuovamente al modo Program, quindi premere tre volte nuovamente al modo di Misura. per accedere

3)

Note: Le impostazioni realizzate con l'Assistente di avvio rapido sono correlate tra di loro e le eventuali modifiche vengono applicate solo dopo avere selezionato Finish (Fine) nell'ultima tappa. Non usare l'assistente di avvio rapido per modificare parametri individualmente: per ulteriori dettagli, consultare i Parametri riportati nella versione integrale delle Istruzioni operative. (Impostare valori personalizzati solo dopo aver concluso l'avvio rapido.) Per ulteriori informazioni su qualsiasi parametro di avvio rapido consultare la versione integrale delle Istruzioni operative.

1. Assistenti 1.1. Avvio rapido 1.1.1. Avvio rapido - Livello


Questo assistente di avvio rapido raggruppa le impostazioni necessarie per applicazioni semplici di misura di livello. a) Premere la freccia DESTRA per attivare il modo PROGRAM e accedere al men livello 1: MAIN MENU (Men principale). b) Premere due volte la freccia DESTRA c) Premere la freccia DESTRA per accedere all'item di men 1.1.1. per attivare l'Assistente di avvio rapido.

1)Vedi il manuale per la Comunicazione SITRANS LUT400 (HART)

7ML19985XU81

SITRANS LUT400 (HART)

Pagina IT-17

d) Premere ogni volta la freccia BASSO per confermare i valori preimpostati e accedere direttamente all'item successivo, o premere la freccia DESTRA per accedere al modo Editazione: la selezione attuale evidenziata. e) Accedere all'item desiderato e premere la per memorizzare la freccia DESTRA modifica, quindi premere la freccia BASSO per continuare. f) L'utente pu premere la freccia ALTO o la freccia SINISTRA

QUICK START LEVEL


TRANSDUCER XRS-5
BACK CANCEL NEXT EDIT

QUICK START LEVEL


TRANSDUCER

NO TRANSDUCER XPS-10 XPS-15 XCT-8 XCT-12

XRS-5

in qualsiasi momento per tornare indietro,

per chiudere l'Assistente.

Italiano

Avvio dell'Assistente di avvio rapido - Livello

Indica il tipo di Assistente scelto per realizzare l'impostazione.

Opzioni
Trasduttore

CANCEL (Annulla), START (Avvio)

Consente di impostare il trasduttore Siemens collegato all'apparecchio..

Opzioni
Funzionamento

NO TRANSDUCER (Senza trasduttore), XRS-5, XPS-10, XPS-15, XCT-8, XCT-12, XPS-30, XPS-40, XLT-30, XLT-60, STH Valore preimpostato: NO TRANSDUCER (Senza trasduttore)

Consente di impostare il tipo di misura (e l'uscita analogica corrispondente) richiesto per l'applicazione.
piano di riferimento sensore distanza punto di calibrazione alto spazio

livello punto di calibrazione basso

range lontano

Modo
LEVEL (Livello) SPACE (Spazio) DISTANCE (Distanza)

Descrizione
* Altezza materiale

Piano di riferimento
Punto di calibrazione basso (livello vuoto) Punto di calibrazione alto (livello pieno) Piano di riferimento sensore (Sensor Reference Point) (segue pagina successiva)

Distanza alla superficie del materiale

Pagina IT-18

SITRANS LUT400 (HART)

7ML19985XU81

(segue dalla pagina precedente)

Modo

Descrizione
NON selezionare. Se il valore di funzionamento OTHER (Altro), l'apparecchio regolato per controllare variazioni di livello ma stato impostato precedentemente nel modo di funzionamento LEVEL (Livello), SPACE (Spazio), o DISTANCE (Distanza). Per avviare l'Assistente di avvio rapido Livello occorre impostare il modo di funzionamento LEVEL, SPACE, o DISTANCE.

Piano di riferimento

OTHER (Altro)

Fonte della temperatura

Fonte della misura di temperatura utilizzata per regolare la velocit del suono.

Italiano

Opzioni

TRANSDUCER (Trasduttore), FIXED TEMPERATURE (Temp. fissa), EXTERNAL TS-3 (TS-3 esterno), AVERAGE OF SENSORS (Media dei sensori) Valore preimpostato: TRANSDUCER (Trasduttore)

Temperatura fissa

Questa funzione utile quando non stato installato un dispositivo di controllo della temperatura.

Valore

Range: da -100.0 a +150.0 C Valore preimpostato: +20.0 C

Questo parametro viene visualizzato solo se stata selezionata FIXED TEMPERATURE (Temp. fissa) per Fonte della temperatura. Unit

Unit di misura sensore.

Opzioni

M, CM, MM, FT, IN Valore preimpostato: M

Nota: In questo esempio, tutti i valori sono espressi in metri (m).


Punto di calibrazione alto

Distanza tra il piano di riferimento sensore e il punto di calibrazione alto: generalmente il livello processo pieno.

Valore

Range: da 0.000 a 60.000 Valore preimpostato: 0.000

Punto di calibrazione basso

Distanza tra il piano di riferimento sensore e il punto di calibrazione basso: generalmente livello processo vuoto.

Valore

Range: da 0.000 a 60.000 Valore preimpostato: 60.000

7ML19985XU81

SITRANS LUT400 (HART)

Pagina IT-19

Velocit di risposta

Regola la risposta di misurazione dell'apparecchio all'interno della portata.

Note:

La velocit di risposta deve essere impostata con l'Assistente di avvio rapido. Eventuali modifiche ai parametri Fill Rate per Minute (2.3.1.), Empty Rate per Minute (2.3.2.), o Damping Filter (2.3.3.) dopo la fine dell'Assistente sostituiranno la Velocit di risposta impostata. Velocit di risposta viene sempre visualizzato in m/minuto.
SLOW (Lento) (0,1 M/MIN) MEDIUM (Medio) (1,0 M/MIN) FAST (Veloce) (10 M/MIN) Valore preimpostato: SLOW (Lento) (0,1 M/MIN)

Opzioni

Italiano

Impostare un valore leggermente superiore alla massima velocit di riempimento o svuotamento (a seconda del valore pi alto). Pi la velocit di risposta bassa, maggiore sar la precisione. Pi alta, maggiori saranno le variazioni del livello. Fine dell'Assistente di avvio rapido - Livello

L'esito dell'avvio rapido risulta positivo solo se vengono applicate tutte le modifiche.

Opzioni

BACK (Indietro), CANCEL (Cancella), FINISH (Fine) (Il display indica 1.1 Quick Start menu (Men Avvio Rapido) se l'Assistente di avvio rapido ha esito positivo o viene annullato. Selezionando CANCEL i valori modificati non vengono memorizzati nell'apparecchio).

Per trasmettere all'apparecchio i valori impostati durante l'avvio rapido e accedere nuovamente al modo Program, premere la freccia BASSO la freccia SINISTRA (Fine). Quindi premere tre volte per accedere nuovamente al modo Measurement.

1.1.2. Avvio rapido - Volume


Questo assistente di avvio rapido raggruppa le impostazioni necessarie per applicazioni con forme di serbatoio standard. a) Premere la freccia DESTRA per attivare il modo PROGRAM e accedere al men livello 1: MAIN MENU (Men principale). b) Premere due volte la freccia DESTRA c) Premere la freccia BASSO avvio rapido Volume. per accedere all'item di men 1.1.1. per attivare l'Assistente di
QUICK START VOLUME
TRANSDUCER XRS-5
BACK CANCEL NEXT EDIT

e la freccia DESTRA

d) Premere ogni volta la freccia BASSO per confermare i valori preimpostati e accedere direttamente all'item successivo, o premere la freccia DESTRA per accedere al modo Editazione: la selezione attuale evidenziata. e) Accedere all'item desiderato e premere la freccia DESTRA f) per memorizzare la modifica, quindi per continuare. in qualsiasi premere la freccia BASSO

QUICK START VOLUME


TRANSDUCER

NO TRANSDUCER XPS-10 XPS-15 XCT-8 XCT-12

XRS-5

L'utente pu premere la freccia ALTO

momento per tornare indietro, o la freccia SINISTRA per chiudere l'Assistente. Pagina IT-20 SITRANS LUT400 (HART) 7ML19985XU81

Avvio dell'Assistente di avvio rapido - Volume

Indica il tipo di Assistente scelto per realizzare l'impostazione.

Opzioni
Trasduttore

CANCEL (Annulla), START (Avvio)

Consente di impostare il trasduttore Siemens collegato all'apparecchio..

Opzioni

NO TRANSDUCER (Senza trasduttore), XRS-5, XPS-10, XPS-15, XCT-8, XCT12, XPS-30, XPS-40, XLT-30, XLT-60, STH Valore preimpostato: NO TRANSDUCER (Senza trasduttore)

Fonte della temperatura

Fonte della misura di temperatura utilizzata per regolare la velocit del suono.

Opzioni

TRANSDUCER (Trasduttore), FIXED TEMPERATURE (Temp. fissa), EXTERNAL TS-3 (TS-3 esterno), AVERAGE OF SENSORS (Media dei sensori) Valore preimpostato: TRANSDUCER (Trasduttore)

Italiano

Temperatura fissa

Questa funzione utile quando non stato installato un dispositivo di controllo della temperatura.

Valore

Range: da -100.0 a +150.0 C Valore preimpostato: +20.0 C

Questo parametro viene visualizzato solo se stata selezionataFIXED TEMPERATURE (Temp. fissa) per Fonte della temperatura. Forma del serbatoio

Definisce la forma del serbatoio e consente di utilizzare SITRANS LUT400 per determinare il volume invece del livello. Selezionando NONE non viene realizzata la conversione del volume. Definire la forma del serbatoio che corrisponde al serbatoio o silo controllato.

Opzioni

NONE, LINEAR, CYLINDER, PARABOLIC BOTTOM, HALF SPHERE BOTTOM, FLAT SLOPED BOTTOM, PARABOLIC ENDS, SPHERE, CONICAL BOTTOM, CURVE TABLE, LINEAR TABLE Valore preimpostato: LINEAR

Selezionando CURVE TABLE o LINEAR TABLE impostare valori per i punti di intervento livello e volume alla fine dell'avvio rapido. (Consultare la versione integrale 2.6.7.Table 1-8 delle Istruzioni operative.) Unit

Unit di misura sensore.

Opzioni

M, CM, MM, FT, IN Valore preimpostato: M

Nota: In questo esempio, tutti i valori sono espressi in metri (m).

7ML19985XU81

SITRANS LUT400 (HART)

Pagina IT-21

Punto di calibrazione alto

Distanza tra il piano di riferimento sensore e il punto di calibrazione alto: generalmente il livello processo pieno.

Valore

Range: da 0.000 a 60.000 Valore preimpostato: 0.000

Punto di calibrazione basso

Distanza tra il piano di riferimento sensore e il punto di calibrazione basso: generalmente livello processo vuoto..

Valore

Range: da 0.000 a 60.000 Valore preimpostato: 60.000

Velocit di risposta

Italiano

Regola la risposta di misurazione dell'apparecchio all'interno della portata.

Note:

La velocit di risposta deve essere impostata con l'Assistente di avvio rapido. Eventuali modifiche ai parametri Fill Rate per Minute (2.3.1.) or Empty Rate per Minute (2.3.2.) dopo la fine dell'Assistente sostituiranno la Velocit di risposta impostata. Velocit di risposta viene sempre visualizzato in m/minuto.
SLOW (Lento) (0,1 M/MIN) MEDIUM (Medio) (1,0 M/MIN) FAST (Veloce) (10 M/MIN) Valore preimpostato: SLOW (Lento) (0,1 M/MIN)

Opzioni

Impostare un valore leggermente superiore alla massima velocit di riempimento o svuotamento (a seconda del valore pi alto). Pi la velocit di risposta bassa, maggiore sar la precisione. Pi alta, maggiori saranno le variazioni del livello. Dimensione A

Altezza del fondo del serbatoio, conico, piramidale, parabolico, sferico o piatto inclinato..

Valore
Dimensione L

Range: da 0.000 a 99.999 Valore preimpostato: 0.000

Lunghezza della sezione cilindrica di un serbatorio orizzontale con estremit paraboliche.

Valore
Unit di volume

Range: da 0.000 a 99.999 Valore preimpostato: 0.000

Definisce le unit di misura del volume.

Opzioni

L, USGAL, IMPGAL, CUM, USER DEFINED * Valore preimpostato: L

* Selezionando USER DEFINED necessario impostare il valore alla fine dell'avvio rapido. Per ulteriori informazioni vedi Parametri 2.6.6.User Defined Unit nella versione integrale delle Istruzioni operative. Pagina IT-22 SITRANS LUT400 (HART) 7ML19985XU81

Volume massimo

Volume massimo del serbatoio. Introdurre il volume del serbatoio corrispondente al piano di riferimento superiore (High Calibration Point). Ad esempio, se il volume massimo del serbatoio 8000 L, introdurre il valore 8000.

Valore

Range: da 0.0 a 9999999 Valore preimpostato: 100.0

Fine dell'Assistente di avvio rapido - Volume

L'esito dell'avvio rapido risulta positivo solo se vengono applicate tutte le modifiche.

Opzioni

BACK (indietro), CANCEL (annulla), FINISH (fine) (il display indica 1.1 Quick Start Wizard menu alla fine dell'Assistente. Selezionando CANCEL i valori modificati non vengono memorizzati nell'apparecchio).

Per trasmettere all'apparecchio i valori impostati durante l'avvio rapido e accedere nuovamente al modo Program, premere la freccia BASSO la freccia SINISTRA (Fine). Quindi premere tre volte per accedere nuovamente al modo Measurement.

Italiano

1.1.3. Avvio rapido - Portata


Questo assistente di avvio rapido raggruppa le impostazioni necessarie per applicazioni semplici di misura di portata. [L'Assistente di avvio Rapido Portata viene indicato sull'interfaccia LUI per il LUT430 (Controllo pompe e Portata), e per il LUT440 (OCM; misura di portata in canale aperto); solo modelli impostati.] a) Premere la freccia DESTRA per attivare il modo PROGRAM e accedere al men livello 1: MAIN MENU (Men principale). b) Premere due volte la freccia DESTRA c) Premere due volte la freccia BASSO l'Assistente di avvio rapido Portata. per accedere all'item di men 1.1.1. e la freccia DESTRA
QUICK START FLOW
TRANSDUCER XRS-5
BACK CANCEL NEXT EDIT

per attivare

d) Premere ogni volta la freccia BASSO per confermare i valori preimpostati e accedere direttamente all'item successivo, o premere la freccia DESTRA per accedere al modo Editazione: la selezione attuale evidenziata. e) Accedere all'item desiderato e premere la per memorizzare la freccia DESTRA modifica, quindi premere la freccia BASSO per continuare. f) L'utente pu premere la freccia ALTO o la freccia SINISTRA

QUICK START FLOW


TRANSDUCER

NO TRANSDUCER XPS-10 XPS-15 XCT-8 XCT-12

XRS-5

in qualsiasi momento per tornare indietro,

per chiudere l'Assistente.

Avvio dell'Assistente di avvio rapido - Portata

Indica il tipo di Assistente scelto per realizzare l'impostazione.

Opzioni

CANCEL (Annulla), START (Avvio)

7ML19985XU81

SITRANS LUT400 (HART)

Pagina IT-23

Trasduttore

Consente di impostare il trasduttore Siemens collegato all'apparecchio..

Opzioni

NO TRANSDUCER (Senza trasduttore), XRS-5, XPS-10, XPS-15, XCT-8, XCT-12, XPS-30, XPS-40, XLT-30, XLT-60, STH Valore preimpostato: NO TRANSDUCER (Senza trasduttore)

Fonte della temperatura

Fonte della misura di temperatura utilizzata per regolare la velocit del suono.

Opzioni

TRANSDUCER (Trasduttore), FIXED TEMPERATURE (Temp. fissa), EXTERNAL TS-3 (TS-3 esterno), AVERAGE OF SENSORS (Media dei sensori) Valore preimpostato: TRANSDUCER (Trasduttore)

Temperatura fissa

Fonte della misura di temperatura utilizzata per regolare la velocit del suono.

Italiano

Valore

Range: da -100.0 a +150.0 C Valore preimpostato: +20.0 C

Questo parametro viene visualizzato solo se stata selezionataFIXED TEMPERATURE (Temp. fissa) per Fonte della temperatura. Elemento primario di misura (Primary Measuring Device)

Definisce il tipo di elemento primario di misura (PMD) utilizzato nell'applicazione.


EXPONENTIAL DEVICES (Elementi esponenziali), RECTANGULAR FLUME BS3680 (Canale rettangolare), ROUND NOSE HORIZONTAL CR. BS-3680 (Stramazzo a parete orizzontale e spigolo arrotondato), TRAPEZOIDAL FLUME BS-3680 (canale trapezoidale), U-FLUME BS-3680 (canale 'U'), FINITE CREST WEIR BS-3680 (stramazzo con parete finita), THIN PLATE RECT. WEIR BS-3680 (stramazzo rettangolare a piastra sottile) , THIN PLATE V-NOTCH WEIR BS-3680 (stramazzo rettangolare a piastra sottile), RECT. WEIR CONTRACTED (stramazzo rettangolare contratto), ROUND PIPE (condotta a sezione circolare), PALMER BOWLUS FLUME (canale Palmer Bowlus), H-FLUME (Canale H), OTHER (Altri)* Valore preimpostato: EXPONENTIAL DEVICES (Elementi esponenziali)

Opzioni

* L'opzione OTHER impostata se l'Assistente di avvio rapido stato previamente lanciato tramite un software compatibile con il protocollo HART (per esempio SIMATIC PDM), e l'apparecchio stato impostato su OFF o UNIVERSAL HEAD VS. FLOW. Se si tratta della prima configurazione PMD pu essere impostato solo su (OFF), o linearizzazione (UNIVERSAL HEAD VS. FLOW; area universale x velocit) tramite tools software HART (SIMATIC PDM, AMS, FC375/475). Consultare la versione integrale delle Istruzioni operative. Metodo di calcolo portata

Definisce il metodo utilizzato per calcolare la portata.

Opzioni
Unit

ABSOLUTE (assoluto), RATIOMETRIC (raziometrico) Valore preimpostato: ABSOLUTE

Unit di misura sensore.

Opzioni

M, CM, MM, FT, IN Valore preimpostato: M

Nota: In questo esempio, tutti i valori sono espressi in metri (m).


Pagina IT-24 SITRANS LUT400 (HART) 7ML19985XU81

Punto di calibrazione alto

Distanza tra il piano di riferimento sensore e il punto di calibrazione alto: generalmente il livello processo pieno.

Valore

Range: da 0.000 a 60.000 Valore preimpostato: 0.000

Punto di calibrazione basso

Distanza tra il piano di riferimento sensore e il punto di calibrazione basso: generalmente livello processo vuoto..

Valore

Range: da 0.000 a 60.000 Valore preimpostato: 60.000

Velocit di risposta

Regola la risposta di misurazione dell'apparecchio all'interno della portata.

Italiano

Note:

La velocit di risposta deve essere impostata con l'Assistente di avvio rapido. Eventuali modifiche ai parametri Fill Rate per Minute (2.3.1.) or Empty Rate per Minute (2.3.2.) dopo la fine dell'Assistente sostituiranno la Velocit di risposta impostata. Velocit di risposta viene sempre visualizzato in m/minuto.
SLOW (Lento) (0,1 M/MIN) MEDIUM (Medio) (1,0 M/MIN) FAST (Veloce) (10 M/MIN) Valore preimpostato: SLOW (Lento) (0,1 M/MIN)

Opzioni

Impostare un valore leggermente superiore alla massima velocit di riempimento o svuotamento (a seconda del valore pi alto). Pi la velocit di risposta bassa, maggiore sar la precisione. Pi alta, maggiori saranno le variazioni del livello. Fattori di calcolo:

Note:

I cinque parametri indicati vengono visualizzati durante l'avvio rapido in base al PMD selezionato sopra. Questi parametri sono utilizzati nella formula di calcolo della portata (consultare la versione integrale delle Istruzioni operative). Esponente portata (PMD = EXPONENTIAL DEVICES; elementi esponenziali)

Esponente utilizzato nella formula di calcolo della portata.

Valore

Range: da -999.000 a 9999.000 Valore preimpostato: 1.550

7ML19985XU81

SITRANS LUT400 (HART)

Pagina IT-25

Fattore K (PMD = EXPONENTIAL DEVICES; elementi esponenziali)

Costante utilizzata nel calcolo della portata per il calcolo assoluto di un elemento esponenziale, esclusivamente.

Valore

Range: da -999.000 a 9999.000 Valore preimpostato: 1.000

Angolo stramazzo (PMD = THIN PLATE V-NOTCH WEIR; stramazzo rettangolare a piastra sottile)

Angolo dello stramazzo impiegato nella formula di calcolo della portata.

Valore Italiano
Pendenza

Range: da 25.000 a 95.000 Valore preimpostato: 25.000

(PMD = TRAPEZOIDAL FLUME (canale trapezoidale) o ROUND PIPE (condotta a sezione circolare)

Pendenza impiegata nella formula di calcolo della portata.

Valore

Range: da -999.000 a 9999.000 Valore preimpostato: 0.000

Coefficente di rugosit (PMD = ROUND PIPE; condotta a sezione circolare)

Coefficente di rugosit impiegato nella formula di calcolo della portata.

Valore

Range: da -999.000 a 9999.000 Valore preimpostato: 0.000

Pagina IT-26

SITRANS LUT400 (HART)

7ML19985XU81

Dimensioni PMD

Note:

Per ogni PMD (elemento primario di misura) esclusi gli Elementi esponenziali, e Altri, necessario impostare fino a quattro dimensioni. Durante l'avvio rapido l'operatore dovr inserire le dimensioni richieste per il PMD selezionato. Verr visualizzato il nome della dimensione corrispondente. PMD selezionato Nome dimensione visualizzato durante l'avvio rapido (men parametri)
APPROACH WIDTH B; larghezza canale d'ingresso (2.15.4.5. OCM Dimension 1) APPROACH WIDTH B; larghezza gola (2.15.4.6. OCM Dimension 2) HUMP HEIGHT P; altezza gradino (2.15.4.7. OCM Dimension 3) THROAT LENGTH L; lunghezza gola (2.15.4.8. OCM Dimension 4)

Canale rettangolare BS-3680

Italiano

Stramazzo a parete orizzontale e spigolo arrotondato BS-3680


CREST WIDTH B; larghezza soglia (2.15.4.5. OCM Dimension 1) CREST HEIGHT P; altezza soglia (2.15.4.6. OCM Dimension 2) CREST LENGTH L; lunghezza soglia (2.15.4.7. OCM Dimension 3)

Canale trapezoidale BS-3680


APPROACH WIDTH B; larghezza canale d'ingresso (2.15.4.5. OCM Dimension 1) APPROACH WIDTH B; larghezza gola (2.15.4.6. OCM Dimension 2) HUMP HEIGHT P; altezza gradino (2.15.4.7. OCM Dimension 3) THROAT LENGTH L; lunghezza gola (2.15.4.8. OCM Dimension 4)

Canale 'U' BS-3680


APPROACH DIAMETER DA; diametro canale d'ingresso (2.15.4.5. OCM Dimension 1) THROAT DIAMETER D; diametro gola (2.15.4.6. OCM Dimension 2) HUMP HEIGHT P; altezza gradino (2.15.4.7. OCM Dimension 3) THROAT LENGTH L; lunghezza gola (2.15.4.8. OCM Dimension 4)

Stramazzo con parete finita BS-3680


CREST WIDTH B; larghezza soglia (2.15.4.5. OCM Dimension 1) CREST HEIGHT P; altezza soglia (2.15.4.6. OCM Dimension 2) CREST LENGTH L; lunghezza soglia (2.15.4.7. OCM Dimension 3)

Stramazzo rettangolare a piastra sottile BS-3680


APPROACH WIDTH B; larghezza canale d'ingresso (2.15.4.5. OCM Dimension 1) CREST WIDTH B; larghezza soglia (2.15.4.6. OCM Dimension 2) CREST HEIGHT P; altezza soglia (2.15.4.7. OCM Dimension 3)

Stramazzo rettangolare contratto


CREST WIDTH B; larghezza soglia (2.15.4.5. OCM Dimension 1) (vedi pagina successiva)

7ML19985XU81

SITRANS LUT400 (HART)

Pagina IT-27

PMD selezionato (segue)

Nome dimensione visualizzato durante l'avvio rapido (men parametri)

Condotta a sezione circolare


THROAT DIAMETER D; diametro interno della condotta (2.15.4.5. OCM Dimension 1)

Canale Palmer Bowlus


MAXIMUM FLUME WIDTH HMAX; massimo battente HMAX (2.15.4.5. OCM Dimension 1)

Canale H
MAXIMUM LISTED HEAD HMAX; massimo battente HMAX (2.15.4.5. OCM Dimension 1)

Italiano

Massimo battente

Massimo valore di livello relativo all'elemento primario (PMD) selezionato.

Valore

Range: da 0.000 a 60.000 Valore preimpostato: 60.000

Offset battente zero

La differenza (positiva o negativa) tra Punto di calibrazione basso e il battente zero (livello se la portata zero).

Valore
Unit di portata

Range: da -60.000 a 60.000 Valore preimpostato: 0.000

Unit di volume utilizzate per indicare la portata totale.

Opzioni

L/S, L/MIN, CUFT/S, CUFT/D, GAL/MIN, GAL/D, IMPGAL/MIN, IMPGAL/D, CUM/H, CUM/D, USER DEFINED (definite dall'utente)* Valore preimpostato: L/S

* Selezionando USER DEFINED necessario impostare il valore alla fine dell'avvio rapido. Per ulteriori informazioni vedi Parametri 2.6.6.User Defined Unit nella versione integrale delle Istruzioni operative. Portata massima a 20 mA

Portata massima.

Valore

Range: da -999 a 9999999 Valore preimpostato: 100

Unit decimali; valore di portata

Numero massimo di unit decimali che pu essere visualizzato.

Opzioni

NO DIGITS (zero cifre), 1 DIGIT (1 cifra), 2 DIGITS, 3 DIGITS Valore preimpostato: NO DIGITS

Limite portata bassa

Disattiva il totalizzatore quando il livello di battente uguale o inferiore al limite.

Valore

Range: da 0.000 a 60.000 Valore preimpostato: 0.000

Pagina IT-28

SITRANS LUT400 (HART)

7ML19985XU81

Fine dell'Assistente di avvio rapido - Portata

L'esito dell'avvio rapido risulta positivo solo se vengono applicate tutte le modifiche.

Opzioni

BACK (indietro), CANCEL (annulla), FINISH (fine) (il display indica 1.1 Quick Start Wizard menu alla fine dell'Assistente. Selezionando CANCEL i valori modificati non vengono memorizzati nell'apparecchio).

Per trasmettere all'apparecchio i valori impostati durante l'avvio rapido e accedere nuovamente al modo Program, premere la frecciaBASSO la freccia SINISTRA (Fine). Quindi premere tre volte per accedere nuovamente al modo Measurement.

Avvertenza: E' vivamente consigliato realizzare un 2.15.2.Auto Zero Head dopo la


fine dell'Assistente per ottenere la massima precisione. Consultare la versione integrale delle Istruzioni operative.

Italiano

1.2. Controllo pompe


Questo assistente di avvio rapido consente di configurare le pompe utilizzate nella vostra applicazione. Assicurarsi di seguire la sequenza di avvio rapido applicabile. a) Premere la freccia DESTRA per attivare il modo PROGRAM e accedere al men livello 1: MAIN MENU (Men principale). b) Premere la freccia DESTRA c) Premere la freccia BASSO per accedere all'item men 1.1. e DESTRA per selezionare il Controllo Pompe.

d) Premere ogni volta la freccia BASSO per confermare i valori preimpostati e accedere direttamente all'item successivo, o premere la freccia DESTRA attuale evidenziata. per accedere al modo Editazione: la selezione per memorizzare la

e) Accedere all'item desiderato e premere la freccia DESTRA modifica, quindi premere la freccia BASSO f) L'utente pu premere la freccia ALTO o la freccia SINISTRA per continuare.

in qualsiasi momento per tornare indietro,

per chiudere l'Assistente.

Avvio dell'Assistente di avvio rapido - Controllo pompa

Indica il tipo di Assistente scelto per realizzare l'impostazione.

Opzioni

CANCEL (Annulla), START (Avvio)

Numero di pompe

Selezionare il numero di pompe utilizzate per il controllo pompe.

Opzioni

NONE (nessuna), 2 Valore preimpostato: NONE

Selezionando NONE il controllo pompe viene disattivato.

7ML19985XU81

SITRANS LUT400 (HART)

Pagina IT-29

Rel pompa 1

Seleziona il rel assegnato alla Pompa 1.

Opzioni
Rel pompa 2

RELAY 2, RELAY 3 Valore preimpostato: RELAY 2

Sola lettura. Definisce automaticamente il rel assegnato alla Pompa 2 in base al rel selezionato per la Pompa 1 nella tappa precedente.

Opzioni (sola lettura)

Se il rel Pompa 1 = RELE' 2, il rel Pompa 2 = RELE' 3 Se il rel Pompa 1 = RELE' 3, il rel Pompa 2 = RELE' 2

Modo controllo pompe

Definisce l'algoritmo di controllo utilizzato per ottenere la commutazione del rel.

Italiano

LUT420 per il controllo di livello:


ALTERNATE DUTY ASSIST, ALTERNATE DUTY BACKUP

LUT430 per il controllo di pompe/portata: Opzioni utilizzabili a seconda del modello


ALTERNATE DUTY ASSIST, ALTERNATE DUTY BACKUP , SERVICE RATIO DUTY ASSIST, SERVICE RATIO DUTY BACKUP , FIXED DUTY ASSIST, FIXED DUTY BACKUP

LUT440 per la misura di portata in canale aperto:


ALTERNATE DUTY ASSIST, ALTERNATE DUTY BACKUP , SERVICE RATIO DUTY ASSIST, SERVICE RATIO DUTY BACKUP , FIXED DUTY ASSIST, FIXED DUTY BACKUP Valore preimpostato (tutti i modelli): ALTERNATE DUTY ASSIST

Rapporto operativo Pompa 1

Seleziona le pompe utilizzate in base alla frequenza e alla durata di utilizzo, non in base alla sequenza di pompaggio.

Valore

Range: da 0 a 255 Valore preimpostato: 1

Questo parametro viene visualizzato solo se stato selezionato un'algoritmo di Rapporto operativo, applicabile al Modo controllo pompe. Rapporto operativo Pompa 2

Seleziona le pompe utilizzate in base alla frequenza e alla durata di utilizzo, non in base alla sequenza di pompaggio.

Valore

Range: da 0 a 255 Valore preimpostato: 1

Questo parametro viene visualizzato solo se stato selezionato un'algoritmo di Rapporto operativo, applicabile al Modo controllo pompe.

Pagina IT-30

SITRANS LUT400 (HART)

7ML19985XU81

Durata di utilizzo rel 2

Definire la durata di utilizzo del rel 2, in ore.

Valore

Range: da 0 a 999999 Valore preimpostato: 0

Utilizzare il valore preimpostato per le pompe nuove, o impostare questo valore per le pompe esistenti con un periodo di utilizzo esistente. (Questo parametro viene visualizzato solo se stato selezionato un'algoritmo di Rapporto operativo, applicabile al Modo controllo pompe.) Durata di utilizzo rel 3

Definire la durata di utilizzo del rel 3, in ore.

Valore

Range: da 0 a 999999 Valore preimpostato: 0

Italiano

Utilizzare il valore preimpostato per le pompe nuove, o impostare questo valore per le pompe esistenti con un periodo di utilizzo esistente. (Questo parametro viene visualizzato solo se stato selezionato un'algoritmo di Rapporto operativo, applicabile al Modo controllo pompe.) Punto di intervento ON Pompa 1

Livello al quale viene attivata la Pompa 1, definito in 2.1.1.Units.

Valore

Range: da 0.000 a 99999.000 Valore preimpostato: 0.000

Punto di intervento ON Pompa 2

Livello al quale viene attivata la Pompa 2, definito in 2.1.1.Units.

Valore

Range: da 0.000 a 99999.000 Valore preimpostato: 0.000

Punto di intervento OFF Pompa 1

Livello al quale viene disattivata la Pompa 1, definito in 2.1.1.Units.

Valore

Range: da 0.000 a 99999.000 Valore preimpostato: 0.000

Punto di intervento OFF Pompa 2

Livello al quale viene disattivata la Pompa 2, definito in 2.1.1.Units.

Valore

Range: da 0.000 a 99999.000 Valore preimpostato: 0.000

Fine dell'Assistente di avvio rapido - Controllo pompa

L'esito dell'avvio rapido risulta positivo solo se vengono applicate tutte le modifiche.

Opzioni

BACK (indietro), CANCEL (annulla), FINISH (fine) (il display indica il men Controllo pompe alla fine dell'Assistente. Selezionando CANCEL i valori modificati non vengono memorizzati nell'apparecchio).

Per trasmettere all'apparecchio i valori impostati durante l'avvio rapido e accedere nuovamente al modo Program, premere la freccia BASSO la freccia SINISTRA 7ML19985XU81 (Fine). Quindi premere due volte per accedere nuovamente al modo Measurement. SITRANS LUT400 (HART) Pagina IT-31

Comunicazione SITRANS LUT400: HART


SITRANS LUT400 pu essere configurato tramite la rete HART usando lo strumento HART Communicator 375/475 della Emerson (per ulteriori informazioni vedi la versione integrale delle Istruzioni operative), o un pacchetto software adatto. E' consigliabile configurare l'apparecchio con SIMATIC Process Device Manager (PDM) di Siemens. Per ulteriori dettagli vedere la guida applicativa Come utilizzare il protocollo HART, scaricabile dalla pagina prodotto del nostro sito internet. Vedi: www.siemens.com/sitransLUT400 rubrica Support e cliccare su Application Guides (Guide all'uso).

Funzionamento remoto
SITRANS LUT400 compatibile con diversi software per il funzionamento attraverso la comunicazione remota: PC con SIMATIC PDM PC con Emerson AMS Device Manager PC con browser web PC con Field Device Tool (FDT) Field Communicator 375/475 (FC375/FC475). Le informazioni basiche circa l'uso di questi software con il SITRANS LUT400 sono descritte qui sotto. Per ulteriori informazioni circa i vari tools consultare le istruzioni operative per la Comunicazione SITRANS LUT400 (HART).

Italiano

SIMATIC PDM
SIMATIC PDM un pacchetto software per la messa in servizio e la manutenzione del SITRANS LUT400 e di altre apparecchiature da campo. Per ulteriori informazioni sul software potete consultare www.siemens.com/simatic-pdm. Consultare la pagina Support del nostro sito internet per ottenere l'ultima versione di SIMATIC PDM, del Service Pack (SP) e del hotfix (HF). Vedere: http://support.automation.siemens.com/WW/ llisapi.dll?func=cslib.csinfo&lang=en&objiD=10806857&subtype=133100

Aggiornare il file EDD (Electronic Device Description)


Il file pu essere scaricato dalla rubrica EDD Device Catalog, sotto Sensors/Level/Echo/Siemens AG/SITRANS LUT400. (Il file EDD garantisce la compatibilit in avanti.) Per facilitare il rilevamento del file descrittivo EDD corretto, necessario assicurasi che i numeri maggiori e minori della revisione del file EDD e del firmware installate siano identici (ad esempio numeri maggiori e minori in caratteri in grassetto: 1.00.00-02). La revisione del file EDD deve corrispondere a quella del software del sistema. Per verificare con PDM, vedi SITRANS LUT400 HART > Identification > Device.

equivalenza tra revisione firmware e revisione EDD

L'installazione di una versione nuova di SIMATIC PDM richiede l'ultima versione del Service Pack (SP) e del hotfix (HF).

Per installare un nuovo file EDD

Vedi www.siemens.com/sitransLUT400 > Support > Software Downloads per scaricare l'ultimo aggiornamento EDD dalla pagina prodotto del nostro sito internet. Memorizzare i file sul vostro computer, ed estrarre il file zip assicurandosi che sia facilmente accessibile. Avviare SIMATIC PDM Manage Device Catalog, accedere al file EDD dezippato e selezionarlo. SITRANS LUT400 (HART) 7ML19985XU81

Pagina IT-32

Impostazione di un apparecchio
1) 2) E' importante disporre dell'ultima versione del file EDD. Il file aggiornabile (vediPer installare un nuovo file EDD). Impostare il SITRANS LUT400 con l'Assistente di avvio rapido adatto per la vostra applicazione.

Accesso agli assistenti di avvio rapido tramite SIMATIC PDM


Note:
Esiste una correlazione tra le diverse impostazioni. Le modifiche vengono applicate cliccando Apply and Transfer dopo l'ultima tappa. Cliccare su BACK (Precedente) per correggere un'impostazione o CANCEL (Annulla) per uscire dal programma di Avvio rapido. Accedere al men Device SIMATIC PDM e selezionare l'Assistente di avvio rapido adatto per la vostra applicazione (Livello, Volume, Volume-Linearizzazione, Portata).

Italiano

Avvio rapido

Uso con AMS Device Manager


AMS Device Manager un pacchetto software per la messa in servizio e la manutenzione del SITRANS LUT400 e di altre apparecchiature da campo. Per ulteriori informazioni sul software: http://www.emersonprocess.com/AMS/.

Electronic Device Description (EDD)


SITRANS LUT400 richiede l'utilizzo del file descrittivo EDD per la versione 10.5 di AMS Device Manager.

Configurazione di un'apparecchio con AMS Device Manager


1) Il file pu essere scaricato dalla rubrica EDD Device Catalog, sotto Sensors/Level/Echo/ Siemens AG/SITRANS LUT400. Consultare la pagina prodotto sul nostro sito internet: www.siemens.com/sitransLUT400, sotto Support->Software Downloads, per ottenere l'ultimo file descrittivo EDD per AMS Device Manager. Memorizzare i file sul vostro computer, ed estrarre il file zip assicurandosi che sia facilmente accessibile. Avviare AMS Device Manager - Add Device Type, accedere al file EDD dezippato e selezionarlo. (L'utente dispone di una Guida di applicazione per apparecchiature HART con AMS Device Manager, scaricabile dal nostro sito internet. Vedi Support.) Device Connection View, fare un clic a destra sull'icona dell'apparecchio e selezionare Scan Device per caricare i parametri dall'apparecchio. Fare doppio clic sull'icona apparecchio per accedere alla finestra di messa in servizio. La finestra di messa in servizio contiene l'identificazione dell'apparechio, e una finestra di navigazione a sinistra. Impostare il SITRANS LUT400 con l'Assistente di avvio rapido adatto per la vostra applicazione.

2) 3) 4) 5) 6)

7ML19985XU81

SITRANS LUT400 (HART)

Pagina IT-33

Funzionamento con FC375/475 (HART)


L'interfaccia di comunicazione FC375/FC475 HART Communicator molto facile da usare e offre compatibilit con molti apparecchi HART, ad esempio SITRANS LUT400. La struttura di men del programmatore FC375/475 molto simile a quella di AMS Device Manager.

Electronic Device Description (EDD)


SITRANS LUT400 richiede il file descrittivo EDD per FC375/FC475.

Configurazione di un apparecchiatura con FC375/FC475


1) Consultare la pagina prodotto sur nostro sito internet www.siemens.com/sitransLUT400, rubrica Support->Software Downloads, per ottenere l'ultima versione del EDD per FC375/ FC475. Memorizzare i file sul vostro computer, ed estrarre il file zip assicurandosi che sia facilmente accessibile. Per installare un nuovo file EDD, attivare il programma Field Communicator Easy Upgrade Utility di Emerson. Impostare il SITRANS LUT400 con l'Assistente di avvio rapido adatto per la vostra applicazione.

2)

Italiano

3) 4)

Uso con la tecnologia FDT (Field Device Tool)


FDT uno standard di interfaccia progettato per la messa in servizio e la manutenzione di apparecchiature di campo, ad esempio SITRANS LUT400. PACTware e Fieldcare sono esempi di software FDT. FDT molto simile a PDM. Per configurare un'apparecchiatura con FDT necessario un DTM (Device Type Manager). Gli strumenti Siemens impiegano SIMATIC PDM e una descrizione dell'apparecchiatura (EDD) per SIMATIC DTM. 1) 2) Per prima cosa necessario installare SITRANS DTM. Il programma scaricabile da: http://www.siemens.com/sitransdtm. Fare clic su Support e quindi Software downloads. Installare la descrizione EDD SITRANS LUT400 (HART) per SITRANS DTM. La lista scaricabile dalla pagina Prodotto del nostro sito internet: Vedi www.siemens.com/ sitransLUT400 rubrica Support e fare clic su Software Downloads.

Configurazione di un apparecchiatura con FDT


L'utente ha la possibilit di scaricare una guida di applicazione dal nostro sito internet, sezione Support.

Servizio e Manutenzione
SITRANS LUT400 non richiede interventi di manutenzione o pulizia in condizioni operative normali.

Riparazione dell'apparecchio ed esclusione di responsabilit


Per ulteriori informazioni consultare la copertina finale di questo manuale.

Pagina IT-34

SITRANS LUT400 (HART)

7ML19985XU81

Cablaggio per aree a rischio esplosione


Per l'installazione in aree a rischio esplosione possibile scegliere fra le seguenti opzioni: Cablaggio a prova di esplosione (Internazionale/Europa) Cablaggio per applicazioni non a rischio incendio (Canada) Per informazioni sulle omologazioni, si consiglia in tutti i casi di consultare la targhetta dell'apparecchiatura.

1. Cablaggio a prova di esplosione (Internazionale/Europa)


SITRANS LUT420 7MLxxxx-xxxxx-xxxx Serial No.: GYZ / B1034567 0518 KCC-REM-S49 Power Rating: 100 230V 15% SITRANSLUT400 50/60 Hz, 36 VA (10 W) Contact Rating: 1A/5A @ 250V , Non-Inductive Operating Temperature: 20C to 50C Enclosure: IP65 / TYPE 4X / NEMA 4X II 3D Ex tc IIIC T70C Dc Ex tc IIIC T70C Dc SIRA 12ATEX9143X IECEx SIR 12.0057X Warning: Possible static hazard. Do not rub or clean on site with a dry cloth. Warning: Do not open when energized or when an explosive dust atmosphere is present. Siemens Milltronics Process Instruments, Peterborough Assembled in Canada with domestic and imported parts SITRANS LUT430 7MLxxxx-xxxxx-xxxx Serial No.: GYZ / B1034567 0518 KCC-REM-S49 Power Rating: 100 230V 15% SITRANSLUT400 50/60 Hz, 36 VA (10 W) Contact Rating: 1A/5A @ 250V , Non-Inductive Operating Temperature: 20C to 50C Enclosure: IP65 / TYPE 4X / NEMA 4X II 3D Ex tc IIIC T70C Dc Ex tc IIIC T70C Dc SIRA 12ATEX9143X IECEx SIR 12.0057X Warning: Possible static hazard. Do not rub or clean on site with a dry cloth. Warning: Do not open when energized or when an explosive dust atmosphere is present. Siemens Milltronics Process Instruments, Peterborough Assembled in Canada with domestic and imported parts SITRANS LUT440 7MLxxxx-xxxxx-xxxx Serial No.: GYZ / B1034567 0518 KCC-REM-S49 Power Rating: 100 230V 15% SITRANSLUT400 50/60 Hz, 36 VA (10 W) Contact Rating: 1A/5A @ 250V , Non-Inductive Operating Temperature: 20C to 50C Enclosure: IP65 / TYPE 4X / NEMA 4X II 3D Ex tc IIIC T70C Dc Ex tc IIIC T70C Dc SIRA 12ATEX9143X IECEx SIR 12.0057X Warning: Possible static hazard. Do not rub or clean on site with a dry cloth. Warning: Do not open when energized or when an explosive dust atmosphere is present. Siemens Milltronics Process Instruments, Peterborough Assembled in Canada with domestic and imported parts

Italiano

SITRANS LUT420 7MLxxxx-xxxxx-xxxx Serial No.: GYZ / B1034567 0518 KCC-REM-S49 Power Rating: 10 32V , 10W SITRANSLUT400 Contact Rating: 1A/5A @ 250V , Non-Inductive Operating Temperature: 20C to 50C Enclosure: IP65 / TYPE 4X / NEMA 4X

SITRANS LUT430 7MLxxxx-xxxxx-xxxx Serial No.: GYZ / B1034567 0518 KCC-REM-S49 Power Rating: 10 32V , 10W SITRANSLUT400 Contact Rating: 1A/5A @ 250V , Non-Inductive Operating Temperature: 20C to 50C Enclosure: IP65 / TYPE 4X / NEMA 4X

SITRANS LUT440 7MLxxxx-xxxxx-xxxx Serial No.: GYZ / B1034567 0518 KCC-REM-S49 Power Rating: 10 32V , 10W SITRANSLUT400 Contact Rating: 1A/5A @ 250V , Non-Inductive Operating Temperature: 20C to 50C Enclosure: IP65 / TYPE 4X / NEMA 4X

II 3D Ex tc IIIC T70C Dc Ex tc IIIC T70C Dc SIRA 12ATEX9143X IECEx SIR 12.0057X Warning: Possible static hazard. Do not rub or clean on site with a dry cloth. Warning: Do not open when energized or when an explosive dust atmosphere is present.

II 3D Ex tc IIIC T70C Dc Ex tc IIIC T70C Dc SIRA 12ATEX9143X IECEx SIR 12.0057X Warning: Possible static hazard. Do not rub or clean on site with a dry cloth. Warning: Do not open when energized or when an explosive dust atmosphere is present.

II 3D Ex tc IIIC T70C Dc Ex tc IIIC T70C Dc SIRA 12ATEX9143X IECEx SIR 12.0057X Warning: Possible static hazard. Do not rub or clean on site with a dry cloth. Warning: Do not open when energized or when an explosive dust atmosphere is present.

Siemens Milltronics Process Instruments, Peterborough Assembled in Canada with domestic and imported parts

Siemens Milltronics Process Instruments, Peterborough Assembled in Canada with domestic and imported parts

Siemens Milltronics Process Instruments, Peterborough Assembled in Canada with domestic and imported parts

Il certificato ATEX scaricabile dalla pagina prodotti del nostro sito www.siemens.com/sitransLUT400 Il certificato IECEx riportato sulla targhetta dell'apparecchio disponibile sul sito internet IECEx. Vedi: http://iecex.iec.ch e cliccare su Ex Equipment Certificates of Conformity poi digitare il numero di certificato IECEx SIR 12.0057X. Per il cablaggio attenersi alle prescrizioni vigenti. Vedi anche e la certificazione ATEX sopraindicata.

7ML19985XU81

SITRANS LUT400 (HART)

Pagina IT-35

2. Cablaggio per applicazioni non a rischio incendio (Canada)


SITRANS LUT420 7MLxxxx-xxxxx-xxxx Serial No.: GYZ / B1034567 Power Rating: 100 230V 15% 50/60 Hz, 36 VA (10 W) Contact Rating: 1A/5A @ 250V , Non-Inductive Operating Temperature: 20C to 50C Enclosure: IP65 / TYPE 4X / NEMA 4X SITRANS LUT430 7MLxxxx-xxxxx-xxxx Serial No.: GYZ / B1034567 Power Rating: 100 230V 15% 50/60 Hz, 36 VA (10 W) Contact Rating: 1A/5A @ 250V , Non-Inductive Operating Temperature: 20C to 50C Enclosure: IP65 / TYPE 4X / NEMA 4X SITRANS LUT440 7MLxxxx-xxxxx-xxxx Serial No.: GYZ / B1034567 Power Rating: 100 230V 15% 50/60 Hz, 36 VA (10 W) Contact Rating: 1A/5A @ 250V , Non-Inductive Operating Temperature: 20C to 50C Enclosure: IP65 / TYPE 4X / NEMA 4X

159134

Class I, Div.2, Gr. A, B, C & D T3C Class II, Div. 2, Gr. F & G Class III Per Drawing A5E03936871

159134

Class I, Div.2, Gr. A, B, C & D T3C Class II, Div. 2, Gr. F & G Class III Per Drawing A5E03936871

159134

Class I, Div.2, Gr. A, B, C & D T3C Class II, Div. 2, Gr. F & G Class III Per Drawing A5E03936871

Siemens Milltronics Process Instruments, Peterborough Assembled in Canada with domestic and imported parts

Siemens Milltronics Process Instruments, Peterborough Assembled in Canada with domestic and imported parts

Siemens Milltronics Process Instruments, Peterborough Assembled in Canada with domestic and imported parts

Italiano

SITRANS LUT420 7MLxxxx-xxxxx-xxxx Serial No.: GYZ / B1034567 Power Rating: 10 32V , 10W Contact Rating: 1A/5A @ 250V , Non-Inductive Operating Temperature: 20C to 50C Enclosure: IP65 / TYPE 4X / NEMA 4X

SITRANS LUT430 7MLxxxx-xxxxx-xxxx Serial No.: GYZ / B1034567 Power Rating: 10 32V , 10W Contact Rating: 1A/5A @ 250V , Non-Inductive Operating Temperature: 20C to 50C Enclosure: IP65 / TYPE 4X / NEMA 4X

SITRANS LUT440 7MLxxxx-xxxxx-xxxx Serial No.: GYZ / B1034567 Power Rating: 10 32V , 10W Contact Rating: 1A/5A @ 250V , Non-Inductive Operating Temperature: 20C to 50C Enclosure: IP65 / TYPE 4X / NEMA 4X

159134

Class I, Div.2, Gr. A, B, C & D T3C Class II, Div. 2, Gr. F & G Class III Per DWG. A5E03936871

159134

Class I, Div.2, Gr. A, B, C & D T3C Class II, Div. 2, Gr. F & G Class III Per DWG. A5E03936871

159134

Class I, Div.2, Gr. A, B, C & D T3C Class II, Div. 2, Gr. F & G Class III Per DWG. A5E03936871

Siemens Milltronics Process Instruments, Peterborough Assembled in Canada with domestic and imported parts

Siemens Milltronics Process Instruments, Peterborough Assembled in Canada with domestic and imported parts

Siemens Milltronics Process Instruments, Peterborough Assembled in Canada with domestic and imported parts

Il diagramma di collegamento A5E03936871 CSA Classe I, Div 2 scaricabile dalla pagina prodotto del nostro sito internet www.siemens.com/sitransLUT400.

Istruzioni specifiche per aree potenzialmente esplosive (Direttiva Europea ATEX 94/9/CE, Allegato II, 1/0/6)
Le istruzioni che seguono riguardano apparecchiature contrassegnate dal numero di certificato SIRA 12ATEX9143X: 1) 2) 3) 4) Per ulteriori informazioni sull'uso e l'assemblaggio, consultare le istruzioni operative. L'apparecchio certificato per l'uso come dispositivo di Categoria 3D. Lapparecchio pu essere installato in atmosfere esplosive zona 22. La temperatura massima della custodia di questo apparecchio T70 C (temperatura ambiente +50 C). Valutare l'impiego di questo apparecchio in base alle normative locali correnti relative a temperature che possono dar luogo all'esplosione di polveri. 5) Questo apparecchio certificado resiste a temperature ambientali tra -20 e +50 C. 6) Gli ingressi-cavi o conduit devono garantire il rispetto dei requisiti della Direttiva 94/9/CE per il Gruppo II, Categoria 3D e garantire il grado di protezione IP della custodia. 7) I cavi devono essere selezionati in conformit con le normative locali ricorrenti, ed essere in grado di sopportare la temperatura massima superficiale della custodia (T70 C). 8) Questa apparecchio non stata definito dispositivo di sicurezza nell'ambito della direttiva 94/9/CE Allegato II, clausola 1,5. 9) L'installazione e l'ispezione di questo apparecchio devono essere eseguite da personale qualificato in conformit con le normative locali correnti (EN 60079-14 e EN 60079-17 in Europa). 10) Le riparazioni di questo dispositivo devono essere eseguite in conformit con i codici di procedura applicabili (EN 60079-19 in Europa).

Pagina IT-36

SITRANS LUT400 (HART)

7ML19985XU81

11) I componenti incorporati nel dispositivo o utilizzati per sostituire altre parti dovranno essere installati da personale qualificato in conformit con le instruzioni contenute nella documentazione fornita dal fabbricante. 12) Per l'intervento manuale, utilizzare il sezionatore fornito nell'installazione del controller associato.

CONDIZIONI SPECIFICHE PER L'USO SICURO


Il suffisso X del numero di certificato si riferisce alle seguenti condizioni operative particolari per garantire la sicurezza. Gli ingressi-cavi o conduit devono garantire il rispetto dei requisiti della Direttiva Europea 94/9/CE per il Gruppo II, Categoria 3D e garantire il grado di protezione IP della custodia. I cavi devono essere selezionati in conformit con le normative locali ricorrenti, ed essere in grado di sopportare la temperatura massima superficiale della custodia (T70 C). La custodia comprende parti non conduttive e pu generare un livello di carica elettrostatica innescabile in particolari condizioni estreme. L'utente deve assicurarsi che l'apparecchio non venga installato in un luogo in cui potrebbe essere soggetto a condizioni esterne (ad esempio vapore alta pression) che potrebbero causare un accumulo di carica elettrostatica sulle superfici non conduttive. E' inoltre importante pulire l'apparecchio solo con un panno umido. L'apparecchio deve essere installato in modo tale da essere protetto dai danni meccanici. Se questo non fosse possibile, l'apparecchio dovr essere protetto con una protezione meccanica adatta per evitare urti o danni. I controllori di livello serie LUT400 (LUT420, LUT430, LUT440) utilizzati con un coperchio per montaggio pannello remoto devono essere installati in una custodia adatta che garantisca la conformit ai requisiti della Direttiva Europea 94/9/CE relativa al Gruppo II, Categoria 3D, con grado di protezione minimo IP65 .

Italiano

7ML19985XU81

SITRANS LUT400 (HART)

Pagina IT-37

Note

Italiano
Pagina IT-38

SITRANS LUT400 (HART)

7ML19985XU81

SITRANS LUT400 (HART) Quick Start Handleiding


Deze handleiding beschrijft de essentile kenmerken en functies van de SITRANS LUT400 (HART 1)). Wij raden u sterk aan de uitgebreide versie van de handleiding aan te schaffen zodat u uw instrument zo optimaal mogelijk kunt gebruiken. De volledige SITRANS LUT400 (HART) gebruikershandleiding, alsmede de communicatie voor SITRANS LUT400 (HART) handleiding, kan worden gedownload van de productpagina van onze website: www.siemens.com/ sitransLUT400. De gedrukte handleiding is verkrijgbaar via uw lokale Siemens vertegenwoordiging. Vragen omtrent de inhoud van deze handleiding kunnen worden gericht aan: Siemens AG Siemens Milltronics Process Instruments 1954 Technology Drive, P .O. Box 4225 Peterborough, Ontario, Canada, K9J 7B1 Email: techpubs.smpi@siemens.com Copyright Siemens AG 2012. Alle rechten voorbehouden
Wij raden aan de geautoriseerde ingebonden handleidingen aan te schaffen, of onze elektronische versies te bekijken, zoals gemaakt en geschreven door Siemens Milltronics Process Instruments. Siemens Milltronics Process Instruments is niet verantwoordelijk voor de inhoud van gedeeltelijke reproducties, van ingebonden of elektronische versies.

Disclaimer
Hoewel we de inhoud van deze gebruikershandleiding hebben geverifieerd aan de omschreven instrumentatie, kunnen desondanks afwijkingen voorkomen. Wij kunnen derhalve niet een volledige overeenstemming garanderen. De inhoud van deze handleiding wordt regelmatig herzien, en correcties worden opgenomen in volgende uitgaven. Wij houden ons aanbevolen voor suggesties ter verbetering. Technische gegevens kunnen worden gewijzigd.

Nederlands

MILLTRONICS is een geregistreerd handelsmerk van Siemens Milltronics Process Instruments.

Technische ondersteuning

Ondersteuning is 24-uur per dag beschikbaar. Ga na de hierna vermelde website om het adres, telefoonnummer en faxnummer van uw lokale Siemens Automation kantoor te vinden: www.siemens.com/automation/partner: Klik op het tabblad Contact om uw productgroep te vinden (bijvoorbeeld: Automation Technology). Klik weer op het productgroep vervolgkeuzemenu om naar de product-familie te gaan. Voor niveau: +Automation Technology > +Sensor Systems >+Process Instrumentation > +Level Measurement > +Continuous. Selecteer Ultrasonic. Voor flow: +Automation Technology > +Sensor Systems >+Process Instrumentation > +Flow Measurement. Selecteer Ultrasonic. Selecteer het land, gevolgd door plaats/regio. Selecteer Technical Support onder Service. Ga voor online technische ondersteuning naar: www.siemens.com/automation/support-request Voer de apparaatnaam in (SITRANS LUT400) of bestelnummer, klik vervolgens op Search en selecteer het juiste producttype. Klik op Next. Voer een sleutelwoord in dat uw vraag omschrijft. Blader vervolgens door de relevante documentatie, of klik op Next om een omschrijving van uw vraag op te sturen naar het technische ondersteuningspersoneel van Siemens. Siemens IA/DT Technical Support Center: telefoonnummer +49 (0)911 895 7222
1)

HART is een geregistreerd handelsmerk van de HART Communications Foundation.

7ML19985XU81

SITRANS LUT400 (HART)

Pagina NL-1

Veiligheidsrichtlijnen
Waarschuwingsmeldingen moeten worden aangehouden om de eigen veiligheid en die van anderen te waarborgen en om het product en de aangesloten apparatuur te beschermen. Deze waarschuwingsmeldingen gaan vergezeld met een verduidelijking van de mate van voorzichtigheid die moet worden aangehouden. WAARSCHUWINGS-symbool Heeft betrekking op een waarschuwingssymbool op het product en betekent dat het niet aanhouden van de noodzakelijke voorzorgsmaatregelen kan leiden tot ernstig of dodelijk letsel en/of aanzienlijke materile schade. WAARSCHUWINGS -symbool wordt gebruikt wanneer er geen corresponderend "voorzichtig" symbool op het product is,betekent dat het niet aanhouden van de noodzakelijke voorzorgsmaatregelen kan leiden tot ernstig of dodelijk letsel en/ of aanzienlijke materile schade. Opmerking: geeft belangrijke informatie aan over het product of dat deel van de gebruikershandleiding.

FCC conformiteit
Alleen voor VS installaties: Federal Communications Commission (FCC) regelgeving
Waarschuwing: wijzigingen of modificaties die niet uitdrukkelijk zijn goedgekeurd door Siemens Milltronics kunnen de toestemming om de apparatuur te bedienen ongeldig maken. Opmerkingen: deze apparatuur is getest, en voldoet aan de grenswaarden voor een Class A digital device, zoals vastgelegd in Part 15 van de FCC regelgeving. Deze grenswaarden zijn vastgelegd voor het bieden van een redelijke bescherming tegen schadelijke interferentie wanneer de apparatuur wordt gebruikt in een commercile omgeving. Deze apparatuur genereert, gebruikt en kan radiofrequentie-energie uitzenden en kan, indien niet gebruikt en genstalleerd conform de gebruikershandleiding schadelijke interferentie veroorzaken voor radiocommunicatie. Gebruik van deze apparatuur in woonomgevingen zal hoogstwaarschijnlijk leiden tot schadelijke interferentie voor radiocommunicatie. In dat geval moet de gebruiker de interferentie voor eigen rekening corrigeren.

Nederlands

Pagina NL-2

SITRANS LUT400 (HART)

7ML19985XU81

SITRANS LUT400
Waarschuwing: de SITRANS LUT400 mag uitsluitend worden gebruikt op de manier zoals beschreven in deze handleiding, anders kan de bescherming die deze apparatuur biedt in gevaar komen. Opmerking: dit product is bedoeld voor gebruik in industrile omgevingen. Gebruik van deze apparatuur in woonomgevingen kan leiden tot interferentie van communicatie op verschillende frequenties. De Siemens SITRANS LUT400 regelaars zijn compact, enkel punt, lang bereik ultrasone regelaars voor continue niveaumeting van vloeistoffen, slurries en stortgoederen, en zeer nauwkeurige bewaking van open kanaal doorstroming. Het programma is compatibel met het hele programma Siemens EchoMax meetversterkers, waardoor een bedrijfsbereik van 0,3 tot 60 meter mogelijk is (afhankelijk van de transducer). SITRANS LUT400 is gekoppeld aan een Local User Interface (LUI) display met achtergrondverlichting en menugestuurde programmering. Het is voorzien van een intutieve interface op web-basis en diverse wizards voor plug en play prestaties. SITRANS LUT400 ondersteunt het HART communicatieprotocol en verschillende softwarepaketten voor communicatie op afstand [SIMATIC PDM, AMS Device Manager, Field Communicator 375/475 en Field Device Tools (FDTs)]. Het is tevens voorzien van onze volgende generatie Sonic Intelligence, waardoor onze positie op het gebied van meetinstrumenten voor de industrie wordt verbeterd terwijl het gebruiksgemak wordt verhoogd.

Specificaties
Voeding
AC uitvoering

Zie voor een compleet overzicht de SITRANS LUT400 (HART) gebruikershandleiding.

Nederlands

100-230 V AC zekering: 5 x 20 mm, traag, 0.25A, 250V

DC uitvoering
10-32 V DC, 10W 1 zekering: 5 x 20 mm, traag, 1.6A, 125V

15%, 50 / 60 Hz, 36 VA (10W)1)

1)

Het maximum opgenomen vermogen weergegeven.

Omgevingscondities
Hoogte
max. 2000 m

Omgevingstemperatuur
-20 tot +50 C (-4 tot +122 F)

Installatieklasse
II

Vervuilingsgraad
4

Relatieve vochtigheid
Geschikt voor buiten (alleen met IP65 / Type 4X / NEMA 4X behuizing)

Goedkeuringen
Opmerkingen: Het typeplaatje van het instrument geeft een opsomming van alle goedkeuringen voor uw instrument. Gebruik de juiste doorvoerafdichtingen om de IP- of NEMA-beschermingsklasse te handhaven. Algemeen CSAUS/C, CE, FM, UL listed, C-TICK Hazardous Niet brandveroorzakend (Canada) CSA Class I, Div. 2, Groups A, B, C, D; Class II, Div 2, Groups F, G; Class III Stofexplosieveilig ATEX II 3D (Internationaal/Europa) Ex tc IIIC T70 C IP65 IECEx SIR 12.0057X SIRA 12ATEX9143X 7ML19985XU81 SITRANS LUT400 (HART) Pagina NL-3

Installatie
Opmerkingen: De installatie mag uitsluitend worden uitgevoerd door gekwalificeerd personeel en in overeenstemming met lokale regelgeving. Dit product is gevoelig voor elektrostatische schokken. Hanteer correcte aardingsprocedures. Alle veldbekabeling moet zijn voorzien van een isolatie geschikt voor tenminste 250 V. Er is een gevaarlijke spanning aanwezig op de klemmen van de sensor tijdens bedrijf. De DC ingangsklemmen moeten worden gevoed uit een bron die galvanisch is gescheiden tussen in- en uitgang, teneinde te voldoen aan de geldende veiligheidsvereisten van IEC 61010-1. Relaiscontacten zijn voor gebruik met apparatuur die geen toegankelijke live onderdelen heeft en bedrading met een isolatie geschikt voor minimaal 250 V. De maximum toegankelijke werkspanning tussen naast elkaar liggende contacten is 250 V. De niet metalen behuizing, bied geen aarding tussen de kabeldoorvoeren. Gebruik aardende bussen en jumpers.

Montage-instructies
Montage-instructies verschillen voor wand-, leiding-, DIN-rail- en apparaten die op een remote display paneel worden gemonteerd. Volg a.u.b. de specifieke instructies voor uw apparaat.

Nederlands

Opmerking: Volg bij het doorvoeren van de kabel door een doorvoer a.u.b. de Kabelleginstructies op pagina 9 voordat u de SITRANS LUT400 monteert. Wand of paneelmontage
Alle configuraties van de SITRANS LUT400 worden verzonden met een rugplaat voor montage. De uitvoering voor paneelmontage wordt geleverd met zowel een LUI display als een deksel.
montage rugplaat

behuizing moet optioneel displaydeksel

clips (2 plaatsen) clips (2 plaatsen)

Opmerking: Bevestigingsmateriaal voor wandmontage niet inbegrepen.

Pagina NL-4

SITRANS LUT400 (HART)

7ML19985XU81

montage rugplaat

behuizing met deksel

clips (2 plaatsen) clips (2 plaatsen)

Opmerking: Bevestigingsmateriaal voor wandmontage niet inbegrepen. achterkant behuizing

sleuven

Nederlands

Wandmontage van de behuizing


1) 2) 3) Markeer en boor vier gaten in het montage-oppervlak voor de vier schroeven (door klant geleverd). Bevestig m.b.v. een schroevendraaier. Breng de sleuven op de achterkant van het apparaat in lijn met de clips op de montageachterplaat. Druk de LUT400 vlak tegen de achterplaat en schuif deze naar beneden op zijn plek.

montageschroefgaten op achterplaat

Let op:
Aanbevolen maat bevestigingsmiddelen: M8 of 5/16" schroef met ring van 17 mm of 5/8" buitendiameter Aanbevolen montage: monteer direct op de wand. Wanneer een alternatief bevestigingsoppervlak wordt gebruikt, dan MOET dit in staat zijn vier maal het gewicht van het apparaat te dragen.

7ML19985XU81

SITRANS LUT400 (HART)

Pagina NL-5

Op afstand gemonteerd deksel

Op afstand display deksel

pakking

4 bevestigingsmiddelen voor paneelmontage meegeleverd

verlengkabel SITRANS LUT400

Nederlands

Montage deksel op afstand


Opmerking: Het op afstand gemonteerde deksel kan op een afstand tot max. 5 m van het apparaat worden gemonteerd, met twee optionele kabelvelengingen (elk 2,5 m lang). Zie de uitgebreide gebruikershandleiding. 1) Gebruik het meegeleverde sjabloon, zaag het benodigde gat uit voor het LUI displaydeksel. Plaats de pakking binnenin het deksel, lijn het uit met de montagegaten. Lijn de achterkant van het displaydeksel op afstand uit met de uitsparing in het paneel. Markeer en boor vier gaten in het montage-oppervlak voor de vier schroeven (meegeleverd). Bevestig met een schroevendraaier en sleutel. Opmerking: Aanbevolen aandraaimoment voor bevestigingsschroeven voor een goede afdichting: 1,1 N m 10 in-lbs

2)

Let op:
Aanbevolen montage: monteer op paneel, tot max. 5 m afstand van het apparaat. Wanneer een alternatief bevestigingsoppervlak wordt gebruikt, dan MOET dit in staat zijn vier maal het gewicht van het apparaat te dragen. Opmerking: Bevestigingsmiddelen inbegrepen: M5 bout, afdichtring, M5 platter ring en moer. Deze zijn nodig om de IP65 classificatie te behouden.

Pagina NL-6

SITRANS LUT400 (HART)

7ML19985XU81

Leidingmontage

montage rugplaat zadelklem

u-bout

behuizing moet optioneel displaydeksel

leiding Opmerking: Bevestigingsmateriaal voor pijpmontage niet inbegrepen.

Montage van de behuizing


1) 2) 3) Bevestig de montagerugplaat op de pijp m.b.v. u-bouten, zadelklemmen, (door klant geleverd) geschikt voor de pijpdiameter. Bevestig de bouten met een sleutel. Niet te strak aandraaien, dan kan de plaat buigen of torderen. Dit kan het vastclippen van de LUT400 op de rugplaat bemoeilijken. Bevestigingsmiddel voor montage rugplaat (zoals omschreven in stap 3 van Wandmontage van de behuizingop pagina 5).

Nederlands

Let op:
Aanbevolen montage: direct op horizontale of verticale pijp. Wanneer een alternatief bevestigingsoppervlak wordt gebruikt, dan MOET dit in staat zijn vier maal het gewicht van het apparaat te dragen. Aanbevolen pijpafmetingen: maximum: 3" pijp, minimum: 3/4" pijp Aanbevolen maat bevestigingsmiddelen: U-bouten: maximum: 3" pijpmaat met M8 of 3/8" schroefdraad minimum: 3/4" pijpmaat met M6 of 1/4" schroefdraad Zeskant moeren: M6 of 1/4" t/m M8 of 3/8" Ring: maximum: 16 mm of 13/16" buitendiameter.

7ML19985XU81

SITRANS LUT400 (HART)

Pagina NL-7

DIN-rail montage
achterkant behuizing

DIN-rail

clips (2)

sleuf

DIN-rail klem

Nederlands

Montage van de behuizing


1) 2) 3) Kantel de bovenkant van de behuizing naar de DIN-rail toe en positioneer deze iets boven de bovenkant van de rail. Beweeg de behuizing naar beneden tegen de DIN-rail om de clips op de achterkant van de behuizing in de bovenkant van de DIN-rail te drukken. Druk het instrument vlak tegen de DIN-rail zodat de DIN-rail-klem aangrijpt en de behuizing stevig op de DIN-rail wordt bevestigd.

Opmerking:
Aanbevolen montage: direct op horizontale DIN-rail. Vereiste DIN-rail afmetingen: TH 35-7.5 of TH 35-15 volgens standaard IEC 60715. De DIN-rail MOET in staat zijn om vier maal het gewicht van de SITRANS LUT400 te ondersteunen.

Verwijder de behuizing
1) Plaats vanaf de voorkant van het apparaat een schroevendraaier in een sleuf onderaan de DIN-rail klem en duw naar beneden. Hierdoor wordt de klem losgeklipt van de onderkant van de DIN-rail. Terwijl u de klem ingedrukt houdt, deze naar boven drukken op de behuizing om de clips van de bovenkant van de rails los te maken.

2)

Pagina NL-8

SITRANS LUT400 (HART)

7ML19985XU81

Voorbereiding voor kabeldoorvoer


Kabels kunnen door een doorvoer worden geleid of de behuizing binnentreden via kabelwartels. Volg stappen 1 t/m 5 hierna om eerst de kabeldoorvoergaten bloot te leggen en vervolgens de stappen uit te voeren voor gebruik met doorvoer, of met kabelwartels. 1) Zorg er voor dat het deksel van de behuizing is gesloten en dat de bevestigingsschroeven zijn geborgd. 2) Plaats de punt van de schroevendraaier in de gleuf van de buitendiameter van de uitbreekopening (zie volgende illustratie). 3) Sla op het uiteinde van de schroevendraaier met de palm van de hand om de uitbreekopening te maken. 4) Draai schroeven los en verwijder het deksel van de behuizing. 5) Verwijder kunststof delen van de uitbreekopeningen uit de behuizing. Zorg er voor dat u de elektronica niet beschadigd met statische elektriciteit, of met het gereedschap dat wordt gebruikt om de uitbreekopeningen te verwijderen.
SITRANS LUT400

kabeldoorvoergaten

uitbreekopening (3 plekken)

groef voor schroevendraaier (3 plekken)

Opmerkingen: Wanneer de uitbreekopening zijn verwijderd, is de toegangsopening 21,4 mm tot 21,6 mm in diameter. M20 wartels (20 mm in diameter) en 1/2" NPT doorvoer (21,3 mm in diameter) passen in dit gat. Wees voorzichtig bij het kiezen van de juiste afdichting voor de toegangsgaten. Een vlakke pakking wordt aanbevolen (i.p.v. O-ring). Wanneer alternatieve kabelwartels worden gebruikt, dan is het de verantwoordelijkheid van de klant om de IP65 beschermingsgraad van de doorvoergaten te handhaven.

Nederlands

Kabel door doorvoer geleid


(vervolg van stap 1 t/m 5 hierboven) 6) Na het voorbereiden voor kabeldoorvoer in stap 1 t/m 5 hierboven, de doorvoer op de behuizing bevestigen. Gebruik alleen fittingen van de juiste maat, goedgekeurd voor waterdichte toepassingen. (Doorvoermaat 1/2" NPT.) 7) Plaats behuizingsdeksel weer en draai de schroeven aan.

Kabel gestript en doorvoer door de wartels


(vervolg van stap 1 t/m 5 hierboven) 6) Na het voorbereiden voor kabeldoorvoer in stap 1 t/m 5 hierboven, wartels losschroeven en stevig tegen de behuizing bevestigen. 7) Voer de kabels door de wartels. Zorg er voor dat de voedingskabel separaat wordt gehouden van de signaalkabels en sluit de kabels aan op de klemmenblokken. 8) Draai de wartels aan om een goede afdichting te vormen. 9) Plaats behuizingsdeksel weer en draai de schroeven aan. Pagina NL-9

7ML19985XU81

SITRANS LUT400 (HART)

Aansluiten
Veiligheidsopmerkingen voor het aansluiten
Raadpleeg voor complete bedradingsinstructies de uitgebreide gebruikershandleiding. Verifieer of alle systeemcomponenten zijn genstalleerd conform de instructies. Sluit alle kabelafschermingen aan op de LUT400 afschermingsklemmen (aangeduid op ). Om verschillende aardpotentialen te voorkomen, het apparaat met het symbool moet u er voor zorgen dat de kabelafschermingen goed zijn geaard. Houd blootgelegde aders van afgeschermde kabels zo kort mogelijk, om ruis in de kabel te reduceren die wordt veroorzaakt door stralingsstrooing en opvangen van ruis.

Aansluiten SITRANS LUT400


WAARSCHUWINGEN:

Controleer het typeplaatje op uw instrument om de toelating te verifiren. Gebruik de juiste doorvoerafdichtingen om de IP- of NEMA-beschermingsklasse te handhaven.
Opmerkingen: Er kunnen aparte kabels en doorvoeren nodig zijn om te voldoen aan de standaard instrumentatie-bekabelingsnormen of elektrische regelgeving. Toegang tot het bedradingscompartiment:

Nederlands

1) 2) 3)

Draai de borgschroeven 1/4 slag los. Til het deksel op naar links op de scharnieren. Het deksel kan open blijven, verbonden via de scharnieren, of het kan worden losgeclipt uit de scharnieren en aan de kant worden gelegd voor toegang tot het bedradingscompartiment. Realiseer alle aansluitingen zoals hierna aangegeven. Wanneer de bedrading is afgerond, plaats dan het deksel van het apparaat weer. Draai de borgschroeven aan.

4) 5) 6)

Pagina NL-10

SITRANS LUT400 (HART)

7ML19985XU81

Bedradingscompartiment
Op de klemmenkaart van de LUT400 kunnen alle ingangen en uitgangen tegelijkertijd worden aangesloten. Klemmenstroken kunnen worden verwijderd voor het gemakkelijker bedraden. Opgelet: Zorg er voor dat de klemmenstroken op de juiste plek worden aangebracht bij het opnieuw plaatsen. Wanneer dit niet gebeurt kan er schade aan het apparaat, of aan aangesloten apparatuur optreden. Opmerking: Aanbevolen aanhaalmoment voor schroefklemmen: 0,56 - 0,79 N m 5 - 7 in-lbs De schroeven niet te strak aandraaien Voeding
TS-3 TRANSDUCER DISCRETE INPUTS INPUT POWER

Voeding

8 1

9 2

10

11 0.25 A, 250 V SLOW BLOW

12 L1

13 L2/N 100 - 230 V

14

DC SYNC

SHLD

WHT

BLK

BLK

WHT

klemmenblokken
mA OUT/HART BATTERY BR2032

batterij

RELAYS

DISPLAY 15 16 RLY 1 17 18 19 20 21 22 23 24

RLY 2

RLY 3 ACT

PAS SERVICE

USBaansluiting

Nederlands

display-kabel

mA HART aansluiting

Voeding
WAARSCHUWINGEN: de DC ingangsklemmen moeten worden gevoed uit een bron die galvanisch is gescheiden tussen in- en uitgang, teneinde te voldoen aan de geldende veiligheidsvereisten van IEC 61010-1. Alle veldbekabeling moet zijn voorzien van een isolatie die geschikt is voor nominale spanningen. Belangrijk! Zorg, voordat u voedingsspanning voor de eerste keer aansluit op de SITRANS LUT400, dat eventueel aangesloten alarm/besturings-apparatuur is uitgeschakeld, totdat een bevredigende werking van het systeem is geverifieerd.

7ML19985XU81

SITRANS LUT400 (HART)

Pagina NL-11

De SITRANS LUT400 is leverbaar in modellen met AC of DC voeding.

AC-voeding
INPUT POWER

DC-voeding
INPUT POWER

12

13

14

12 10 - 32 V

13

14

L1 L2/N 100 - 230 V

AC: 100-230 V AC 15%, 50/60 Hz, 36 VA (10W)

DC: 10-32 V DC, 10W

Opmerking: Zorg er voor dat het apparaat is aangesloten op een betrouwbare aarde. 1) 2) 3) Voor het aansluiten op de voeding, kabelmantel strippen over een lengte van ca. 70 mm (2.75") vanaf het einde van de kabel, en voer de draden door de wartel 1). Sluit de draden aan op de klemmen zoals aangegeven: de polariteit wordt aangegeven onder het klemmenblok. Aard het instrument conform de lokale regelgeving. Opmerkingen voor AC voedingsaansluitingen: De apparatuur moet zijn beschermd door een zekering van 15 A, of een veiligheidsautomaat in de elektrische installatie van het gebouw. Een veiligheidsautomaat, of schakelaar in de installatie van het gebouw, gemarkeerd als de UIT-schakelaar, moet zich in de buurt bevinden van de apparatuur en binnen handbereik van de operator zijn.

Nederlands
1)

De accu
De SITRANS LUT400 wordt gevoed met n accu genstalleerd. (Zie illustratie op pagina 11 voor de locatie van de accu.) De accu (BR2032) heeft een verwachte levensduur van tien jaar en wordt benvloed door de omgevingstemperatuur. Wanneer de LUT400 geen voedingsspanning krijgt, dan blijft de accu de real time klok van het instrument voeden totdat de voedingsspanning weer is hersteld. (Zie de uitgebreide gebruikershandleiding voor het vervangen van de accu.)

Wanneer de kabel door een doorvoer wordt gevoerd, gebruik dan uitsluitend wartels van de goede maat die zijn goedgekeurd voor waterdichte toepassingen.

Pagina NL-12

SITRANS LUT400 (HART)

7ML19985XU81

Sensoren
Waarschuwing: Er is een gevaarlijke spanning aanwezig op de klemmen van de sensor tijdens bedrijf. Voer de sensorkabel door een geaarde metalen doorvoer, gescheiden van andere bekabeling (behalve TS-3 temperatuursensor bekabeling, indien van toepassing). Opmerkingen: Gebruik niet de coaxiaalkabel met SITRANS LUT400 vanwege de hoge spanning op de afscherming van de coaxiaalkabel Sluit de LUT400 afscherming en witte sensoraders niet op elkaar aan; sluit ze aan op aparte klemmen Negeer oudere sensorhandleidingen die dit wel aanbevelen
TS-3 TRANSDUCER

SHLD

BLK

WHT

Relais
Relaiscontacten worden weergegeven in de afgevallen stand. Alle relais kunnen worden geconfigureerd als positieve- of negatieve logica (zie parameter 2.8.11.Relay Logic in de uitgebreide gebruikershandleiding).
RELAYS

BLK

WHT

Nederlands

15

16 RLY 1

17

18

19

20

21

RLY 2

RLY 3

Voedingsspanningsstoring Relais 2, 3 zijn NO. Relais 1 kan worden bedraad als NO of NC. In geval van uitval van voedingsspanning, keren relais terug naar hun normale toestand.

Nominale waarden relais:


n Form C (NO of NC) relais (relais 1), 1A bij 250 V AC, niet inductief, 3A bij 30 V DC twee Form A (NO) relais (relais 2,3), 5A bij 250 V AC, niet inductief, 3A bij 30 V DC

7ML19985XU81

SITRANS LUT400 (HART)

Pagina NL-13

Aansluiten van HART mA uitgang (HART)


mA OUT/HART

Voor ACTIEVE HARTaansluiting (met gebruik van de in LUT400 gentegreerde voeding), sluit klemmen 22 en 23 aan.

22

23

24

ACT

PAS

Voor PASSIEVE HARTaansluiting (met gebruik van externe voeding), sluit klemmen 23 en 24 aan.

Raadpleeg voor meer informatie de mA uitgangsparameters (2.5.Current Output) in de parametersectie van de uitgebreide gebruikershandleiding.

Aansluiten bij installaties in explosiegevaarlijke zones


Zie pagina 35.

Inbedrijfstelling SITRANS LUT400


Wij raden aan dat u uw applicatie in de onderstaande volgorde configureert: Voer eerst de juiste Quick Start Wizard uit voor uw toepassing (Level, Volume, Flow). Configureer vervolgens pompen m.b.v. de Pump Control Wizard (indien van toepassing). Tot slot, configureer alarmen, of andere regelingen, totaaltellers en monsternemers, waarbij wordt gerefereerd aan de betreffende parameters (zie uitgebreide gebruikershandleiding). Er zijn twee manieren om toegang te krijgen tot de Quick Start Wizards: lokaal (Zie Quick Start Wizards via LUIop pagina 17) vanaf een locatie op afstand (Zie Andere Quick Start Wizards:op pagina 17)

Nederlands

Activeren van de SITRANS LUT400


Inschakelen van het instrument. SITRANS LUT400 start automatisch op in de meet modus. Een transitiescherm toont eerst het Siemens logo en vervolgens de huidige firmware-revisie van de LUI terwijl de eerste meting wordt verwerkt. De eerste keer dat het apparaat wordt geconfigureerd, vraagt het systeem u een taal te kiezen (Engels, Duits, Frans, Spaans of Chinees). De juiste datum en tijd moeten worden ingesteld voordat het instrument wordt geconfigureerd. (Stel voor het wijzigen parameter 2.14.Date and Time in. Zie de uitgebreide gebruikershandleiding voor details.)

Pagina NL-14

SITRANS LUT400 (HART)

7ML19985XU81

Het LCD Meetmodus weergave: normaal bedrijf 1) Niveau


1 10
SITRANS LUT400 LEVEL

2
1/2

4 5

18.91
1

6
2

5 SAVING PARAMETERS

Doorstroming
1 10 11 12
SITRANS LUT400 FLOW HEAD
1 2 3

3
1/2
L/S M

4 5 6

500313..
60.000

DAILY TOTAL,...

999999.99

5 SAVING PARAMETERS

1 tag 2 meetwaarde (niveau, ruimte, afstand, volume, doorstroming of opvoerhoogte) 3 waarde wordt weergegeven [Primary Variable (PV)=1 of 2, Secondary Variable (SV)=2 of 2] 4 toggle indicator1) voor PV of SV 5 eenheden 6 bargraph geeft niveau aan 7 het secundaire gebiedt geeft de geconfigureerde relais aan (links) en discrete ingangen (rechts) 8 tekstgebied toont statusmeldingen 9 instrument statusindicator 10 geselecteerde (primaire) sensormodus: niveau, ruimte, afstand, volume, opvoerhoogte, of doorstroming 11 secundaire sensormodus = opvoerhoogte wanneer de primaire sensormodus = doorstroming 12 totalizer waarden: toont afwisselend de dag-totalizer en de doorlopende totalizer

Nederlands

Storing aanwezig
0 LOE

8 tekstgebied toont een foutcode en een foutmelding 9 service nodig pictogram verschijnt

PROGRAM modus display Navigatieweergave

Een zichtbare schuifbalk geeft aan dat menu / TRANSDUCER 2.1.6 parameterde menulijst te lang is om alle items in naam parameterNO TRANSDUCER nummer XRS-5 n keer weer te geven. XPS-10 item De hoogte van de balk in het menu XPS-15 band huidige XCT-8 geeft de lengte van de menulijst aan: item XCT-12 XPS-30 Een hogere balk geeft minder items schuifbalk aan. De positie van de of balk geeft bij benadering de positie aan van het huidige item in de lijst. Een balk halverwege het menu geeft aan dat het huidige item halverwege de lijst is. Een hogere balk betekent minder items.

huidige

1)

Druk op

of

om te schakelen.

7ML19985XU81

SITRANS LUT400 (HART)

Pagina NL-15

Parameterweergave
menu waarin zicht de parameter bevindt parameternaam
SENSOR
UNITS SENSOR MODE SENSOR MODE SECONDARY

Bewerken weergave
2.1.6

parameternummer parameterwaarde/ selectie

TRANSDUCER
NO TRANSDUCER XPS-10 XPS-15 XCT-8 XCT-12 XPS-30

2.1.6

XRS-5

TRANSDUCER

FREQUENCY

XRS-5

PROGRAM modus
Druk op de lokale knoppen op het instrument op om de programmeermodus te activeren , , , . en menu-niveau 1 te openen. Blader door het menu m.b.v.

Selecteren van een vermelde optie


a) Navigeer naar de gewenste parameter. b) Druk op de pijl RECHTS gemarkeerd. om de Edit modus te openen De huidige selectie is

c) Blader naar een nieuwe selectie. d) Druk op de pijl RECHTS om deze te accepteren Het display keert terug naar de weergave van de parameters en toont de nieuwe selectie.

Nederlands

Wijzigen van een numerieke waarde


a) Navigeer naar de gewenste parameter, bijv., Low Calibration Point (2.2.1.). b) Indien geselecteerd, dan wordt de huidige waarde weergegeven. c) Druk op pijl RECHTS om de Edit modus te openen. De cursorpositie wordt gemarkeerd. d) Gebruik de pijl LINKS en RECHTS om de cursor te verplaatsen naar de positie die u wilt wijzigen. e) Gebruik bij een gemarkeerde positie de pijl OMHOOG respectievelijk verhogen of verlagen van de positie. f) en OMLAAG voor het en
LOW CALIB. PT. 2.2.1

ESC

+ _ _ _ _ _ _ 2. 0 M

OK

Gebruik wanneer de decimale punt is geselecteerd, de pijl OMHOOG OMLAAG om de decimale punt te verschuiven.

g) Druk om te annuleren zonder de wijzigingen op te slaan continu op de pijl LINKS totdat ESC is gemarkeerd. Druk weer op de pijl LINKS om af te sluiten zonder wijzigingen op te slaan. Druk anders, wanneer de nieuwe parameterwaarde correct is, continu op de pijl RECHTS totdat OK wordt gemarkeerd.

h) Druk op de pijl RECHTS om de nieuwe waarde te accepteren. Het display keert terug naar de weergave van de parameters en toont de nieuwe selectie. Bekijk de waarden nauwkeurig. Pagina NL-16 SITRANS LUT400 (HART) 7ML19985XU81

Quick Start Wizards


Gebruik voor het configureren van de SITRANS LUT400 voor niveau-, volume- (standaard tankvormen) of doorstromingstoepassingen de Quick Start Wizards via LUI zoals hierna beschreven. Een wizard voor toepassingen met complexere tankvormen is leverbaar via SIMATIC PDM. Zie

Quick Start Wizards via SIMATIC PDMop pagina 33.

Andere Quick Start Wizards:


Andere Quick Start Wizards die gebruik maken van verschillende softwarepakketten zijn eveneens leverbaar 1): SIMATIC PDM (HART) AMS (HART) FC375/475 (HART) FDTs (HART) Parameter configuratieformulieren (voor het vastleggen van data voorafgaande aan het configureren van het instrument via de Quick Start Wizard) zijn beschikbaar op onze website. Ga naar www.siemens.com/sitransLUT400 > Support > Application Guides.

Quick Start Wizards via LUI


1) 2) Druk op in de programmamodus te activeren. Kies Wizards (1.), Quick Start (1.1.) en vervolgens de juiste quick start: QS Level (1.1.1.), QS Volume (1.1.2.), of QS Flow (1.1.3.). [De QS Flow wizard wordt alleen getoond op LUI voor LUT430 (pomp en doorstroming), en LUT440 (OCM) geconfigureerde modellen.] Volg de stappen en kies dan Finish om de Quick Start parameterwijzigingen op te slaan en terug te keren naar het programmeermenu, druk vervolgens drie keer op om terug te keren naar de meetmodus. Opmerkingen: De Quick Start wizard instellingen zijn onderling gerelateerd en wijzigingen worden pas toegepast wanneer u op Finish kiest in de laatste stap. Gebruik de Quick Start wizard niet voor het wijzigen van individuele parameters: zie in plaats daarvan de parametersectie in de uitgebreide gebruikershandleiding. (Voer alleen een aanpassing uit voor uw toepassing nadat de Quick Start is afgerond.). Zie voor meer informatie over wizard parameters de uitgebreide gebruikershandleiding.

Nederlands

3)

1. Wizards 1.1. Quick Start 1.1.1. QS Level


Gebruik deze wizard voor het configureren van eenvoudige niveaubewakingstoepassingen. a) Druk op pijl RECHTS om de PROGRAMMER modus te activeren en open menu niveau 1: MAIN MENU. b) Druk op pijl RECHTS twee maal om naar menu item 1.1.1. te gaan c) Druk op pijl RECHTS om het QS-niveau te openen.
1)

Zie Communicatie voor SITRANS LUT400 (HART) handleiding

7ML19985XU81

SITRANS LUT400 (HART)

Pagina NL-17

d) Druk bij elke stap op de pijl OMLAAG om de standaard waarden te accepteren en direct naar het volgende item te gaan, of pijl RECHTS om de Edit modus te openen: De huidige selectie wordt gemarkeerd. e) Blader naar het gewenste item en druk op de pijl RECHTS om de wijziging op te slaan, om druk vervolgens op pijl OMLAAG verder te gaan. f)

QUICK START LEVEL


TRANSDUCER XRS-5
BACK CANCEL NEXT EDIT

QUICK START LEVEL


TRANSDUCER

NO TRANSDUCER XPS-10 XPS-15 XCT-8 XCT-12

XRS-5

Op elk gewenst moment kunt u op de pijl OMHOOG op de pijl LINKS om de wizard te annuleren. Start van QS Level Wizard

drukken om terug te gaan, of

Toont het type Wizard dat uitgevoerd moet worden.

Opties
Antenne

CANCEL, START

Geeft de Siemens transducer aan die is aangesloten op het device.

Opties Nederlands
Bedrijf

NO TRANSDUCER, XRS-5, XPS-10, XPS-15, XCT-8, XCT-12, XPS-30, XPS-40, XLT-30, XLT-60, STH Standaard: NO TRANSDUCER

Stelt het type meting in (en de corresponderende mA uitgang) benodigd voor de toepassing.
Sensor referentiepunt High Calibration Point Ruimte

Afstand

Niveau Low Calibration Point

Ver weg bereik

Modus
Niveau Ruimte Afstand

Omschrijving
* Hoogte van materiaal Afstand tot materiaaloppervlak

reference point (referentiepunt)


Low Calibration Point (proces leeg niveau) High Calibration Point (proces vol niveau) Sensor referentiepunt (vervolgd op volgende pagina)

Pagina NL-18

SITRANS LUT400 (HART)

7ML19985XU81

(vervolg van vorige pagina)

Modus

Omschrijving
NIET selecteren. Wanneer de bedrijfwaarde wordt weergegeven als OTHER, dan is het instrument geconfigureerd als niveaubewakingsinstrument, maar het is eerder ingesteld op een modus anders dan LEVEL, SPACE, of DISTANCE. De bedrijfsmodus moet zijn ingesteld op LEVEL, SPACE, of DISTANCE om verder te gaan met de QS Level Wizard.

reference point (referentiepunt)

andere

Temperatuurbron

Bron van de temperatuurmeetwaarde die wordt gebruikt voor het instellen van de snelheid van het geluid.

Opties

TRANSDUCER, FIXED TEMPERATURE, EXTERNAL TS-3, AVERAGE OF SENSORS Standaard: Antenne

Fixed Temperature

Gebruik deze functie wanneer een temperatuurmeetinstrument niet wordt gebruikt.

Nederlands

Waarde

Bereik: -100.0 tot +150.0 C Standaard: +20.0 C

Deze parameter wordt alleen weergegeven wanneer FIXED TEMPERATURE is gekozen voor Temperatuurbron. Eenheden

Sensor meeteenheiden.

Opties

M, CM, MM, FT, IN Standaard: M

Opmerking: Voor dit voorbeeld zijn alle waarden in meters (m).


High Calibration Point

Afstand van sensorreferentiepunt tot High (Hoog) kalibratiepunt: meestel het proces vol niveau.

Waarde

Bereik: 0.000 ... 60.000 Standaard: 0.000

Low Calibration Point

Afstand van sensorreferentiepunt tot Low (Laag) kalibratiepunt: meestal het proces leeg niveau.

Waarde

Bereik: 0.000 ... 60.000 Standaard: 60.000

7ML19985XU81

SITRANS LUT400 (HART)

Pagina NL-19

Responsiesnelheid

Stelt de reactiesnelheid in van het apparaat op wijzigingen in de meetwaarden in het doelbereik.

Opmerkingen:

De Response Rate kan alleen worden ingesteld via de Quick Start Wizard en eventuele Fill Rate per Minute (2.3.1.), Empty Rate per Minute (2.3.2.), of Damping Filter (2.3.3.) parameters na de afronding van de wizard zullen de instellingen van de Response Rate vervangen. Responsiesnelheid wordt altijd weergegeven in m/minute.
Langzaam (0,1 m/min) Normaal (1.0 m/min) Snel (10 m/min) Standaard: Langzaam (0,1 m/min)

Opties

Gebruik een instelling die net iets sneller is dan de maximale vul- of leegsnelheid (welke maar groter is). Langzamere instellingen bieden een hogere nauwkeurigheid, snellere instellingen laten meer snellere niveaufluctuaties toe. Einde van QS Level Wizard

Om de QS succesvol te laten zijn, moeten alle wijzigingen worden toegepast.

Opties Nederlands

BACK, CANCEL, FINISH (Display keert terug naar 1.1 Quick Start menu wanneer de Quick Start met succes wordt afgerond, of geannuleerd. Wanneer CANCEL wordt geselecteerd, worden er geen wijzigingen naar het apparaat geschreven.)

Voor het overdragen van de Quick Start waarden naar het apparaat en om vervolgens terug te keren naar het Program menu, moet u op de pijl OMLAAG (Finish). Druk vervolgens drie maal op pijl LINKS drukken naar de Meetmodus. om terug te keren

1.1.2. QS Volume
Gebruik deze wizard voor het configureren van volume-toepassingen, waarbij standaard tankvormen worden gebruikt. a) Druk op pijl RECHTS om de PROGRAMMER modus te activeren en open menu niveau 1: MAIN MENU. b) Druk op pijl RECHTS twee maal om naar menu item 1.1.1. te gaan c) Druk op de pijl OMLAAG en pijl RECHTS om QS Volume te openen.
QUICK START VOLUME
TRANSDUCER XRS-5
BACK CANCEL NEXT EDIT

d) Druk bij elke stap op de pijl OMLAAG om de standaard waarden te accepteren en direct naar het volgende item te gaan, of pijl RECHTS om de Edit modus te openen: De huidige selectie wordt gemarkeerd. e) Blader naar het gewenste item en druk op de pijl RECHTS om de wijziging op te slaan, druk vervolgens op pijl OMLAAG f) om verder te gaan. om de Op elk gewenst moment kunt u op de pijl OMHOOG drukken om terug te gaan, of op de pijl LINKS wizard te annuleren. Pagina NL-20 SITRANS LUT400 (HART)

QUICK START VOLUME


TRANSDUCER

NO TRANSDUCER XPS-10 XPS-15 XCT-8 XCT-12

XRS-5

7ML19985XU81

Begin van QS Volume Wizard

Toont het type Wizard dat uitgevoerd moet worden.

Opties
Antenne

CANCEL, START

Geeft de Siemens transducer aan die is aangesloten op het device.

Opties
Temperatuurbron

NO TRANSDUCER, XRS-5, XPS-10, XPS-15, XCT-8, XCT-12, XPS-30, XPS-40, XLT-30, XLT-60, STH Standaard: NO TRANSDUCER

Bron van de temperatuurmeetwaarde die wordt gebruikt voor het instellen van de snelheid van het geluid.

Opties

TRANSDUCER, FIXED TEMPERATURE, EXTERNAL TS-3, AVGERAGE OF SENSORS Standaard: Antenne

Vaste temperatuur

Gebruik deze functie wanneer een temperatuurmeetinstrument niet wordt gebruikt.

Waarde

Bereik: -100.0 tot +150.0 C Standaard: +20.0 C

Deze parameter wordt alleen weergegeven wanneer FIXED TEMPERATURE is gekozen voor Temperatuurbron. Vorm van de tank

Nederlands

Definieert de tankvorm en stelt de SITRANS LUT400 in staat het volume te berekenen in plaats van het niveau. Wanneer NONE wordt gekozen, vindt er geen volumeconversie plaats. Selecteer de tankvorm die overeenkomt met die van de bewaakte tank of reservoir.

Opties

NONE, LINEAR, CYLINDER, PARABOLIC BOTTOM, HALF SPHERE BOTTOM, FLAT SLOPED BOTTOM, PARABOLIC ENDS, SPHERE, CONICAL BOTTOM, CURVE TABLE, LINEAR TABLE Standaard: LINEAR

Wanneer CURVE TABLE of LINEAR TABLE wordt geselecteerd, voer dan de waarden in voor niveau en volume breakpoints na het afronden van de wizard. (Zie parameter 2.6.7.Table 1-8 in de uitgebreide gebruikershandleiding.) eenheden

Sensor meeteenheiden.

Opties

M, CM, MM, FT, IN Standaard: M

Opmerking: Voor dit voorbeeld zijn alle waarden in meters (m).

7ML19985XU81

SITRANS LUT400 (HART)

Pagina NL-21

High Calibration Point

Afstand van sensorreferentiepunt tot High (Hoog) kalibratiepunt: meestel het proces vol niveau.

Waarde

Bereik: 0.000 ... 60.000 Standaard: 0.000

Low Calibration Point

Afstand van sensorreferentiepunt tot Low (Laag) kalibratiepunt: meestal het proces leeg niveau.

Waarde

Bereik: 0.000 ... 60.000 Standaard: 60.000

Responsiesnelheid

Stelt de reactiesnelheid in van het apparaat op wijzigingen in de meetwaarden in het doelbereik.

Opmerkingen:

Response Rate kan alleen worden ingesteld m.b.v. de Quick Start Wizard, en eventuele wijzigingen gemaakt voor de Fill Rate per Minute (2.3.1.) of Empty Rate per Minute (2.3.2.) parameters na afronding van de wizard zullen de instelling van de Response Rate vervangen. Responsiesnelheid wordt altijd weergegeven in m/minute.
Langzaam (0,1 m/min) Normaal (1.0 m/min) Snel (10 m/min) Standaard: Langzaam (0,1 m/min)

Nederlands

Opties

Gebruik een instelling die net iets sneller is dan de maximale vul- of leegsnelheid (welke maar groter is). Langzamere instellingen bieden een hogere nauwkeurigheid, snellere instellingen laten meer niveaufluctuaties toe. Afmeting A

De hoogte van de tankbodem wanneer de bodem conisch, pyramidevormig, parabolisch, sferisch of vlak-hellend is..

Waarde
Dimension L

Bereik: 0.000 ... 99.999 Standaard: 0.000

Lengte van het cilindrische deel van een horizontaal parabolisch uiteinde van een tank.

Waarde
Volume Units

Bereik: 0.000 ... 99.999 Standaard: 0.000

Bepaalt de meeteenheden van het volume.

Opties

L, USGAL, IMPGAL, CUM, USER DEFINED * Standaard: L

*Wanneer de USER DEFINED optie wordt gekozen, dan moet de waarde worden ingesteld na het afronden van de wizard. Zie parameter 2.6.6.User Defined Unit in de uitgebreide gebruikershandleiding. Pagina NL-22 SITRANS LUT400 (HART) 7ML19985XU81

Maximum volume

Het maximum volume van de tank. Voer het tankvolume in overeenkomstig het High Calibration punt. Wanneer bijvoorbeeld uw maximale tankvolume 8000 l is, voer dan een waarde in van 8000.

Waarde

Bereik: 0.0 ... 9999999 Standaard: 100.0

Einde van QS Volume Wizard

Om de QS succesvol te laten zijn, moeten alle wijzigingen worden toegepast.

Opties

BACK, CANCEL, FINISH (Display keert terug naar 1.1 Quick Start menu wanneer de Quick Start succesvol wordt afgerond of geannuleerd. Wanneer CANCEL wordt geselecteerd, worden er geen wijzigingen naar het apparaat geschreven.)

Voor het overdragen van de Quick Start waarden naar het apparaat en om vervolgens terug te keren naar het Program menu, moet u op de pijl OMLAAG drukken (Finish). Druk vervolgens drie maal op pijl LINKS naar de Meetmodus. om terug te keren

1.1.3. QS Flow
Gebruik deze wizard voor het configureren van eenvoudige flow-toepassingen. [Alleen zichtbaar op LUT430 (Pomp en Flow) en LUT440 (OCM) geconfigureerde modellen.] a) Druk op pijl RECHTS om de PROGRAMMER modus te activeren en open menu niveau 1: MAIN MENU. b) Druk op pijl RECHTS twee maal om naar menu item 1.1.1. te gaan en pijl RECHTS om QS Flow te openen.
QUICK START FLOW
TRANSDUCER XRS-5
BACK CANCEL NEXT EDIT

Nederlands

c) Druk twee maal op pijl OMLAAG

d) Druk bij elke stap op de pijl OMLAAG om de standaard waarden te accepteren en direct naar het volgende item te gaan, of pijl RECHTS om de Edit modus te openen: De huidige selectie wordt gemarkeerd. e) Blader naar het gewenste item en druk op de pijl RECHTS om de wijziging op te slaan, om druk vervolgens op pijl OMLAAG verder te gaan. f)

QUICK START FLOW


TRANSDUCER

NO TRANSDUCER XPS-10 XPS-15 XCT-8 XCT-12

XRS-5

Op elk gewenst moment kunt u op de pijl OMHOOG op de pijl LINKS om de wizard te annuleren. Start van QS Flow Wizard

drukken om terug te gaan, of

Toont het type Wizard dat uitgevoerd moet worden.

Opties
7ML19985XU81

CANCEL, START

SITRANS LUT400 (HART)

Pagina NL-23

Antenne

Geeft de Siemens transducer aan die is aangesloten op het device.

Opties
Temperatuurbron

NO TRANSDUCER, XRS-5, XPS-10, XPS-15, XCT-8, XCT-12, XPS-30, XPS-40, XLT-30, XLT-60, STH Standaard: NO TRANSDUCER

Bron van de temperatuurmeetwaarde die wordt gebruikt voor het instellen van de snelheid van het geluid.

Opties

TRANSDUCER, FIXED TEMPERATURE, EXTERNAL TS-3, AVGERAGE OF SENSORS Standaard: Antenne

Vaste temperatuur

Bron van de temperatuurmeetwaarde die wordt gebruikt voor het instellen van de snelheid van het geluid.

Waarde

Bereik: -100.0 tot +150.0 C Standaard: +20.0 C

Deze parameter wordt alleen weergegeven wanneer FIXED TEMPERATURE is gekozen voor Temperatuurbron. Primary Measuring Device

Nederlands

Definieert het primaire meetinstrument (PMD) dat gebruikt moet worden in de toepassing.
EXPONENTIAL DEVICES, RECTANGULAR FLUME BS-3680, ROUND NOSE HORIZONTAL CR. BS-3680, TRAPEZOIDAL FLUME BS-3680, U-FLUME BS3680, FINITE CREST WEIR BS-3680, THIN PLATE RECT. WEIR BS-3680, THIN PLATE V-NOTCH WEIR BS-3680, RECT. WEIR CONTRACTED, ROUND PIPE, PALMER BOWLUS FLUME, H-FLUME, OTHER* Standaard: EXPONENTIAL DEVICES

Opties

* Optie wordt ingesteld op OTHER wanneer de wizard eerder is gedraaid via HART software tool (zoals SIMATIC PDM) en het apparaat was ingesteld op OFF of UNIVERSAL HEAD VS. FLOW. Wanneer dit de initile configuratie is, dan kan de PMD alleen worden ingesteld op geen berekening (OFF), of op linearisatie (UNIVERSAL HEAD VS. FLOW) via HART softwaretools (SIMATIC PDM, AMS, FC375/475). Zie de uitgebreide gebruikershandleiding. Method of Flow Calculation

Stelt de methode van de doorstromingsberekening in.

Opties
eenheden

ABSOLUTE, RATIOMETRIC Standaard: ABSOLUTE

Sensor meeteenheiden.

Opties

M, CM, MM, FT, IN Standaard: M

Opmerking: Ten behoeve van dit voorbeeld zijn alle waarden in meters (m).
Pagina NL-24 SITRANS LUT400 (HART) 7ML19985XU81

High Calibration Point

Afstand van sensorreferentiepunt tot High (Hoog) kalibratiepunt: meestel het proces vol niveau.

Waarde

Bereik: 0.000 ... 60.000 Standaard: 0.000

Low Calibration Point

Afstand van sensorreferentiepunt tot Low (Laag) kalibratiepunt: meestal het proces leeg niveau.

Waarde

Bereik: 0.000 ... 60.000 Standaard: 60.000

Responsiesnelheid

Stelt de reactiesnelheid in van het apparaat op wijzigingen in de meetwaarden in het doelbereik.

Opmerkingen:

Response Rate kan alleen worden ingesteld m.b.v. de Quick Start Wizard, en eventuele wijzigingen gemaakt voor de Fill Rate per Minute (2.3.1.) of Empty Rate per Minute (2.3.2.) parameters na afronding van de wizard zullen de instelling van de Response Rate vervangen. Responsiesnelheid wordt altijd weergegeven in m/minute.
Langzaam (0,1 m/min)

Nederlands

Opties

Normaal (1.0 m/min) Snel (10 m/min) Standaard: Langzaam (0,1 m/min)

Gebruik een instelling die net iets sneller is dan de maximale vul- of leegsnelheid (welke maar groter is). Langzamere instellingen bieden een hogere nauwkeurigheid, snellere instellingen laten meer niveaufluctuaties toe. Berekeningsfactoren:

Opmerkingen:

De volgende vijf parameters worden weergegeven in de wizard op basis van de PMD zoals hierboven geselecteerd. Deze parameters worden gebruikt in de flow berekeningsformule (zie de uitgebreide bedieningshandleiding voor details). Flow Exponent (PMD = EXPONENTIAL DEVICES)

De exponent voor de flow-berekeningsformule.

Waarde

Bereik: -999.000 ... 9999.000 Standaard: 1.550

7ML19985XU81

SITRANS LUT400 (HART)

Pagina NL-25

K Factor (PMD = EXPONENTIAL DEVICES)

De constante die wordt gebruikt in de flow-berekingsformule voor absolute berekening van uitsluitend een exponentieel apparaat.

Waarde
V-Notch Angle

Bereik: -999.000 ... 9999.000 Standaard: 1.000

(PMD = THIN PLATE V-NOTCH WEIR)

De hoek van het gebruikte V-schot die wordt gebruikt in de flow-berekeningsformule.

Waarde
Helling

Bereik: 25.000 ... 95.000 Standaard: 25.000

(PMD = TRAPEZOIDAL FLUME of ROUND PIPE)

De flow-helling die wordt gebruikt in de flow-berekeningsformule.

Waarde

Bereik: -999.000 ... 9999.000 Standaard: 0.000

Roughness Coefficient (PMD = ROUND PIPE)

Nederlands

De flow ruwheidscofficient die wordt gebruikt in de flow-berekeningsformule.

Waarde

Bereik: -999.000 ... 9999.000 Standaard: 0.000

Pagina NL-26

SITRANS LUT400 (HART)

7ML19985XU81

PMD afmetingen

Opmerkingen:

Voor elke PMD m.u.v. Exponential Devices, en Andere, moet u tot vier afmetingen invullen. In de wizard, wordt u gevraagd na elke afmeting die nodig is voor de geselecteerde PMD, en de betreffende PMD dimensie-naam wordt weergegeven. PMD geselecteerd Wizard dimensie naam (parameter menu referentie)
APPROACH WIDTH B (2.15.4.5. OCM Dimension 1) THROAT WIDTH B (2.15.4.6. OCM Dimension 2) HUMP HEIGHT P (2.15.4.7. OCM Dimension 3) THROAT LENGTH L (2.15.4.8. OCM Dimension 4)

Rechthoekige goot BS-3680

Ronde neus horizontaal schot BS-3680


CREST WIDTH B (2.15.4.5. OCM Dimension 1) CREST HEIGHT P (2.15.4.6. OCM Dimension 2) CREST LENGTH L (2.15.4.7. OCM Dimension 3)

Trapezodale goot BS-3680


APPROACH WIDTH B (2.15.4.5. OCM Dimension 1)

Nederlands

THROAT WIDTH B (2.15.4.6. OCM Dimension 2) HUMP HEIGHT P (2.15.4.7. OCM Dimension 3) THROAT LENGTH L (2.15.4.8. OCM Dimension 4)

U-goot BS-3680
APPROACH DIAMETER DA (2.15.4.5. OCM Dimension 1) THROAT DIAMETER D (2.15.4.6. OCM Dimension 2) HUMP HEIGHT P (2.15.4.7. OCM Dimension 3) THROAT LENGTH L (2.15.4.8. OCM Dimension 4)

Overstortschot met vaste lengte BS-3680


CREST WIDTH B (2.15.4.5. OCM Dimension 1) CREST HEIGHT P (2.15.4.6. OCM Dimension 2) CREST LENGTH L (2.15.4.7. OCM Dimension 3)

Rechthoekig schot BS-3680


APPROACH WIDTH B (2.15.4.5. OCM Dimension 1) CREST WIDTH B (2.15.4.6. OCM Dimension 2) CREST HEIGHT P (2.15.4.7. OCM Dimension 3)

Rechthoekig schot vernauwd


CREST WIDTH B (2.15.4.5. OCM Dimension 1) (vervolgd op volgende pagina)

7ML19985XU81

SITRANS LUT400 (HART)

Pagina NL-27

PMD geselecteerd (vervolg) Ronde pijp

Wizard dimensie naam (parameter menu referentie)

PIPE INSIDE DIAMETER D (2.15.4.5. OCM Dimension 1)

Palmer Bowlus goot


MAXIMUM FLUME WIDTH HMAX (2.15.4.5. OCM Dimension 1)

H-goot
MAXIMUM LISTED HEAD HMAX (2.15.4.5. OCM Dimension 1)

Maximum opstuwhoogte

De maximale niveauwaarde bij de PMD.

Waarde

Bereik: 0.000 ... 60.000 Standaard: 60.000

Nul opstuwhoogte Offset

Het verschil (positief of negatief) tussen Low Calibration Point en nul opstuwhoogte (niveau bij nul doorstroming).

Waarde
Flowrate Units

Bereik: -60.000 ... 60.000 Standaard: 0.000

Nederlands

De volume-eenheden worden gebruikt voor het weergeven van de totale flow.

Opties

L/S, L/MIN, CUFT/S, CUFT/D, GAL/MIN, GAL/D, IMPGAL/MIN, IMPGAL/D, CUM/H, CUM/D, USER DEFINED * Standaard: L/S

*Wanneer de USER DEFINED optie wordt gekozen, dan moet de waarde worden ingesteld na het afronden van de wizard. Zie parameter 2.6.6.User Defined Unit in de uitgebreide gebruikershandleiding. Maximale doorstroming bij 20 mA

De maximale flowrate.

Waarde

Bereik: -999 ... 9999999 Standaard: 100

Flowrate decimaal

Het maximum aantal decimalen dat weergegeven moet worden.

Opties
Low Flow Cutoff

NO DIGITS, 1 DIGIT, 2 DIGITS, 3 DIGITS Standaard: NO DIGITS

Elimineert totalizer activiteit voor opstuwhoogtes op of onder de lekstroomwaarde.

Waarde

Bereik: 0.000 ... 60.000 Standaard: 0.000

Pagina NL-28

SITRANS LUT400 (HART)

7ML19985XU81

End of QS Flow Wizard

Om de QS succesvol te laten zijn, moeten alle wijzigingen worden toegepast.

Opties

BACK, CANCEL, FINISH (Display keert terug naar 1.1.1 Quick Start menu wanneer de Quick Start succesvol wordt afgerond of geannuleerd. Wanneer CANCEL wordt geselecteerd, worden er geen wijzigingen naar het apparaat geschreven.)

Voor het overdragen van de Quick Start waarden naar het apparaat en om vervolgens terug te keren naar het Program menu, moet u op de pijl OMLAAG drukken (Finish). Druk vervolgens drie maal op pijl LINKS om terug te keren naar de Meetmodus.

Opmerking: Het wordt sterk aanbevolen dat een 2.15.2.Auto Zero Head wordt
uitgevoerd na afronding van de wizard om de beste nauwkeurigheid te waarborgen. Zie de uitgebreide gebruikershandleiding voor details.

1.2. Pompbesturing
Gebruik deze wizard voor het configureren van pompen wanneer deze in uw toepassing worden gebruikt. Zorg er voor dat u eerst de van toepassing zijnde Quick Start Wizard afrond. a) Druk op pijl RECHTS om de PROGRAMMER modus te activeren en open menu niveau 1: MAIN MENU. b) Druk op de PIJL naar rechts c) Druk op de pijl OMLAAG om naar menu item 1.1 te gaan. en pijl RECHTS om Pompbesturing te openen.

Nederlands

d) Druk bij elke stap op de pijl OMLAAG om de standaard waarden te accepteren en direct naar het volgende item te gaan, of pijl RECHTS om de Edit modus te openen: De huidige selectie is gemarkeerd. om de wijziging op te

e) Blader naar het gewenste item en druk op de pijl RECHTS slaan, druk vervolgens op pijl OMLAAG f)

om verder te gaan. drukken om terug te gaan, of

Op elk gewenst moment kunt u op de pijl OMHOOG op de pijl LINKS om de wizard te annuleren. Start van de Wizard - Pump Control

Toont het type Wizard dat uitgevoerd moet worden.

Opties

CANCEL, START

Number of Pumps

Selecteer het aantal pompen dat gebruikt moet worden bij de pompbesturing.

Opties

NONE, 2 Standaard: NONE

Indien ingesteld op NONE, dan is de pompbesturing uitgeschakeld. 7ML19985XU81 SITRANS LUT400 (HART) Pagina NL-29

Relay Pump 1

Selecteert het relais dat is toegekend aan Pomp 1.

Opties
Relay Pump 2

RELAY 2, RELAY 3 Standaard: Relais 2

View only. Stelt automatisch het relais in dat is toegekend aan Pomp 2 op basis van het relais dat is geselecteerd voor Pomp 1 in de vorige stap.

Options (alleen bekijken)

Wanneer Relay Pump 1 = RELAY 2, dan Relay Pump 2 = RELAY 3 Wanneer Relay Pump 1 = RELAY 3, dan Relay Pump 2 = RELAY 2

Pump Control Mode

Stelt het besturingsalgoritme in dat wordt gebruikt om het relais te schakelen.

LUT420 niveauregelaar:
ALTERNATE DUTY ASSIST, ALTERNATE DUTY BACKUP

LUT430 pomp- en flowregelaar: Opties DUTY ASSIST, SERVICE RATIO DUTY BACKUP , FIXED DUTY ASSIST, FIXED ondersteund DUTY BACKUP per model LUT440 zeer nauwkeurige open kanaal monitor: Nederlands
ALTERNATE DUTY ASSIST, ALTERNATE DUTY BACKUP , SERVICE RATIO DUTY ASSIST, SERVICE RATIO DUTY BACKUP , FIXED DUTY ASSIST, FIXED DUTY BACKUP Standaard (alle modellen): ALTERNATE DUTY ASSIST ALTERNATE DUTY ASSIST, ALTERNATE DUTY BACKUP , SERVICE RATIO

Service Ratio Pump 1

Selecteert pompgebruik op basis van de RUN time ratio in plaats van laatst gebruikt.

Waarde

Bereik: 0 ... 255 Standaard: 1

Deze parameter verschijnt alleen wanneer een Service Ratio algoritme is geselecteerd voor Pump Control Mode. Service Ratio Pump 2

Selecteert pompgebruik gebaseerd op de RUN tijd-ratio i.p.v. laatst gebruikt.

Waarde

Bereik: 0 ... 255 Standaard: 1

Deze parameter verschijnt alleen wanneer een Service Ratio algoritme is geselecteerd voor Pump Control Mode.

Pagina NL-30

SITRANS LUT400 (HART)

7ML19985XU81

Run Time Relay 2

Stel de hoeveelheid tijd in die pomprelais 2 heeft gedraaid, in uren.

Waarde

Bereik: 0 ... 999999 Standaard: 0

Gebruik de standaard waarde voor nieuwe pompen, of stel deze waarde in voor bestaande pompen met geaccumuleerde run-time. Deze parameter verschijnt alleen wanneer een Service Ratio algoritme is geselecteerd voor Pump Control Mode.) Run Time Relay 3

Stel de hoeveelheid tijd in die pomprelais 3 heeft gedraaid, in uren.

Waarde

Bereik: 0 ... 999999 Standaard: 0

Gebruik de standaard waarde voor nieuwe pompen, of stel deze waarde in voor bestaande pompen met geaccumuleerde run-time. Deze parameter verschijnt alleen wanneer een Service Ratio algoritme is geselecteerd voor Pump Control Mode.) ON Setpoint Pump 1

Het niveau waarbij pomp 1 inschakelt, gedefinieerd in 2.1.1.Units.

Waarde

Bereik: 0.000 ... 99999.000 Standaard: 0.000

ON Setpoint Pump 2

Het niveau waarbij pomp 2 inschakelt, gedefinieerd in 2.1.1.Units.

Waarde

Bereik: 0.000 ... 99999.000 Standaard: 0.000

Nederlands

OFF Setpoint Pump 1

Het niveau waarbij pomp 1 uitschakelt, gedefinieerd in 2.1.1.Units.

Waarde

Bereik: 0.000 ... 99999.000 Standaard: 0.000

OFF Setpoint Pump 2

Het niveau waarbij pomp 2 uitschakelt, gedefinieerd in 2.1.1.Units.

Waarde

Bereik: 0.000 ... 99999.000 Standaard: 0.000

Einde van Wizard - Pump Control

Om de Wizard succesvol te laten zijn, moeten alle wijzigingen worden toegepast.

Opties

BACK, CANCEL, FINISH (Display keert terug naar het pompbesturingsmenu wanneer de Wizard succesvol wordt afgerond, of geannuleerd. Wanneer CANCEL wordt geselecteerd, worden er geen wijzigingen naar het apparaat geschreven.)

Voor het overdragen van de waarden naar het apparaat en om vervolgens terug te keren naar het Program menu, moet u op de pijl OMLAAG drukken vervolgens twee maal op pijl LINKS (Finish). Druk om terug te keren naar de Meetmodus.

7ML19985XU81

SITRANS LUT400 (HART)

Pagina NL-31

SITRANS LUT400 Communicatie: HART


De SITRANS LUT400 kan worden geconfigureerd via het HART network m.b.v. de HART Communicator 375/475 van Emerson (zie de uitgebreide gebruikershandleiding voor details), of een softwarepakket. Het aanbevolen softwarepakket is de SIMATIC Process Device Manager (PDM) van Siemens. Details worden beschreven in een toepassingsgids Werken met HART, welke kan worden gedownload vanaf de productpagina van onze website. Ga naar: www.siemens.com/ sitransLUT400 onder Support en klik op Application Guides.

Bediening op afstand
SITRANS LUT400 ondersteunt verschillende softwaretools voor bediening via communicatie op afstand: PC waarop SIMATIC PDM draait PC waarop Emerson AMS Device Manager draait PC waarop een web browser draait PC waarop een Field Device Tool (FDT) draait Field Communicator 375/475 (FC375/FC475). Hieronder vindt u de basis informatie die nodig is om deze gereedschappen te gebruiken met uw SITRANS LUT400. Meer informatie over elke tool kunt u vinden in de Communicatie voor SITRANS LUT400 (HART) handleiding.

SIMATIC PDM
SIMATIC PDM is een softwarepakket dat word gebruikt voor het in bedrijf stellen en onderhouden van de SITRANS LUT400 en andere procesapparaten. Meer informatie kunt u vinden op www.siemens.com/simatic-pdm. Controleer de support pagina van onze website om er zeker van te zijn dat u beschikt over de meest recente versie van SIMATIC PDM, het meest recente Service Pack (SP) en de meest recente hot fix (HF). Ga naar: http://support.automation.siemens.com/WW/ llisapi.dll?func=cslib.csinfo&lang=en&objiD=10806857&subtype=133100

Nederlands

Updaten van de Electronic Device Description (EDD)


U kunt de EDD vinden in de Device Catalogus, onder Sensors/Level/Echo/Siemens AG/ SITRANS LUT400. (De EDD is geschreven met het oog op forward compatibiliteit.) Als richtlijn voor het lokaliseren van de juiste EDD, de major en minor nummers moeten overeenstemmen tussen de EDD revisie en de Firmware revisie in het device (bijv. major en minor nummers in vette tekst: 1.00.00-02).
De EDD-revisie moet overeenkomen met de Firmwarerevisie in het apparaat. Om dit te checken in de PDM, ga naar SITRANS LUT400 HART > Identification > Device.

overeenkomende Firmware en EDD revisies

Voor het installeren van een nieuwe versie van SIMATIC PDM is het meest recente Service Pack (SP) en de meest recente hot fix (HF) nodig.

Installeren van een nieuwe EDD

Ga naar www.siemens.com/sitransLUT400 > Support > Software Downloads voor het downloaden van de meest recente EDD van de productpagina van onze website. Sla de bestanden op uw PC op en pak de gezipte bestanden uit in een gemakkelijk toegankelijke map. Start SIMATIC PDM Manage Device Catalog, blader naar en selecteer de map welke het uitgepakte EDD bestand bevat. SITRANS LUT400 (HART) 7ML19985XU81

Pagina NL-32

Configureren van een nieuw apparaat


1) 2) Verifieer of u beschikt over de meest recente EDD, en update deze indien nodig (zieInstalleren van een nieuwe EDD, hierboven). Configureer de SITRANS LUT400 m.b.v. de Quick Start Wizard voor uw toepassing.

Quick Start Wizards via SIMATIC PDM


Opmerkingen:
De Quick Start wizard instellingen zijn onderling gerelateerd en wijzigingen worden pas toegepast wanneer u op Apply and Transfer klikt aan het einde van stap . Klik op BACK om terug te gaan en een instelling te wijzigen of kies Cancel om de Quick Start te verlaten. Open het Device menu in SIMATIC PDM en selecteer de Quick Start voor uw toepassing (Level, Volume, Volume-Linearization, Flow).

Quick Starts

Nederlands

Bediening via AMS Device Manager


AMS Device Manager is een softwarepakket dat is ontworpen voor het inbedrijf stellen en onderhouden van de SITRANS LUT400 en andere procesinstrumenten. U kunt meer details vinden op: http://www.emersonprocess.com/AMS/.

Electronic Device Description (EDD)


SITRANS LUT400 vereist de EDD voor AMS Device Manager versie 10.5.

Configureren van een nieuw apparaat via AMS Device Manager


1) U kunt de EDD vinden in de Device Catalogus, onder Sensors/Level/Echo/Siemens AG/ SITRANS LUT400. Controleer de productpagina van onze website: www.siemens.com/sitransLUT400, onder Support->Software Downloads, om er zeker van te zijn dat u de meest recente versie heeft van de EDD voor AMS Device Manager. Sla de bestanden op uw computer op en pak het zip-bestand uit in een gemakkelijk toegankelijke map. Start AMS Device Manager Add Device Type, blader naar het uitgepakte EDD-bestand en selecteer het. (Een Applicatiehandleiding voor het instellen van HART apparaten met de AMS Device Manager kunt u downloaden vanaf de productpagina van onze website onder Support) In Device Connection View, klik rechts op het instrumentpictogram en selecteer Scan Device voor het uploaden van parameters van het instrument. Dubbelklik op het instrument pictogram op het startscherm te openen. Het startscherm toont de instrument-identificatiedetails en een navigatievenster aan de linkerkant van het scherm. Configureer de SITRANS LUT400 m.b.v. de juiste Quick Start Wizard voor uw toepassing.

2) 3)

4) 5)

6)

7ML19985XU81

SITRANS LUT400 (HART)

Pagina NL-33

Bediening via FC375/475 (HART)


De FC375/FC475 HART Communicator is een handheld communicatie-instrument dat gemakkelijk is in het gebruik en biedt een universele ondersteuning voor andere HARTinstrumenten zoals de SITRANS LUT400. De menustructuur van de FC375/475 lijkt sterk op die van de AMS Device Manager.

Electronic Device Description (EDD)


SITRANS LUT400 vereist de EDD voor FC375/FC475.

Configureren van een nieuw instrument m.b.v. FC375/FC475


1) Controleer de productpagina van onze website op www.siemens.com/sitransLUT400, onder Support->Software Downloads, om er zeker van te zijn dat u beschikt over de meest recente versie van de EDD voor FC375/FC475. Sla de bestanden op uw computer op en pak het zip-bestand uit in een gemakkelijk toegankelijke map. Draai voor het installeren van de EDD, Emersons Field Communicator Easy Upgrade Utility. Configureer de SITRANS LUT400 m.b.v. de Quick Start Wizard voor uw toepassing.

2) 3) 4)

Bediening via FDT (Field Device Tool)


FDT is een standaard die wordt gebruikt in diverse softwarepakketten die zijn bedoeld voor de inbedrijfstelling en het onderhoud van veldapparaten zoals de SITRANS LUT400. Twee in de handel verkrijgbare FDT's zijn PACTware en Fieldcare.

Nederlands

FDT lijkt sterk op PDM. Voor het configureren van een veldinstrument via FDT, heeft u de DTM (Device Type Manager) voor het instrument nodig. Siemens instruments gebruikt SITRANS DTM en een instrument-EDD geschreven voor SITRANS DTM. 1) Installeer eerst SITRANS DTM op uw systeem. U kunt deze downloaden van: http://www.siemens.com/sitransdtm. Klik op Support en ga vervolgens naar Software downloads. Installeer de SITRANS LUT400 (HART) EDD voor SITRANS DTM. Deze kan worden gedownload vanaf de productpagina van onze website: Ga naar www.siemens.com/ sitransLUT400 onder Support en klik op Software Downloads.

2)

Configureren van een nieuw apparaat via FDT


Een Application Guide kan worden gedownload vanaf de productpagina van onze website onder Support.

Service en onderhoud

De SITRANS LUT400 vereist geen onderhoud of reiniging onder normale bedrijfsomstandigheden.

Reparatie van het instrument en uitsluiting van aansprakelijkheid


Raadpleeg voor gedetailleerde informatie de binnenkant van de achterflap.

Pagina NL-34

SITRANS LUT400 (HART)

7ML19985XU81

Bedrading configuraties voor installaties in explosiegevaarlijke gebieden


De volgende bedradingsopties zijn beschikbaar voor installaties in gevaarlijke gebieden: Stofexplosieveilige bekabeling (Internationaal/Europa) Niet brandveroorzakende bekabeling (Canada) Controleer in alle gevallen, het apparaatlabel op uw instrument en verifieer het type goedkeuring.

1. Stofexplosieveilige bekabeling (Internationaal/Europa)


SITRANS LUT420 7MLxxxx-xxxxx-xxxx Serial No.: GYZ / B1034567 0518 KCC-REM-S49 Power Rating: 100 230V 15% SITRANSLUT400 50/60 Hz, 36 VA (10 W) Contact Rating: 1A/5A @ 250V , Non-Inductive Operating Temperature: 20C to 50C Enclosure: IP65 / TYPE 4X / NEMA 4X II 3D Ex tc IIIC T70C Dc Ex tc IIIC T70C Dc SIRA 12ATEX9143X IECEx SIR 12.0057X Warning: Possible static hazard. Do not rub or clean on site with a dry cloth. Warning: Do not open when energized or when an explosive dust atmosphere is present. Siemens Milltronics Process Instruments, Peterborough Assembled in Canada with domestic and imported parts SITRANS LUT430 7MLxxxx-xxxxx-xxxx Serial No.: GYZ / B1034567 0518 KCC-REM-S49 Power Rating: 100 230V 15% SITRANSLUT400 50/60 Hz, 36 VA (10 W) Contact Rating: 1A/5A @ 250V , Non-Inductive Operating Temperature: 20C to 50C Enclosure: IP65 / TYPE 4X / NEMA 4X II 3D Ex tc IIIC T70C Dc Ex tc IIIC T70C Dc SIRA 12ATEX9143X IECEx SIR 12.0057X Warning: Possible static hazard. Do not rub or clean on site with a dry cloth. Warning: Do not open when energized or when an explosive dust atmosphere is present. Siemens Milltronics Process Instruments, Peterborough Assembled in Canada with domestic and imported parts SITRANS LUT440 7MLxxxx-xxxxx-xxxx Serial No.: GYZ / B1034567 0518 KCC-REM-S49 Power Rating: 100 230V 15% SITRANSLUT400 50/60 Hz, 36 VA (10 W) Contact Rating: 1A/5A @ 250V , Non-Inductive Operating Temperature: 20C to 50C Enclosure: IP65 / TYPE 4X / NEMA 4X II 3D Ex tc IIIC T70C Dc Ex tc IIIC T70C Dc SIRA 12ATEX9143X IECEx SIR 12.0057X Warning: Possible static hazard. Do not rub or clean on site with a dry cloth. Warning: Do not open when energized or when an explosive dust atmosphere is present. Siemens Milltronics Process Instruments, Peterborough Assembled in Canada with domestic and imported parts

SITRANS LUT420 7MLxxxx-xxxxx-xxxx Serial No.: GYZ / B1034567 0518 KCC-REM-S49 Power Rating: 10 32V , 10W SITRANSLUT400 Contact Rating: 1A/5A @ 250V , Non-Inductive Operating Temperature: 20C to 50C Enclosure: IP65 / TYPE 4X / NEMA 4X

SITRANS LUT430 7MLxxxx-xxxxx-xxxx Serial No.: GYZ / B1034567 0518 KCC-REM-S49 Power Rating: 10 32V , 10W SITRANSLUT400 Contact Rating: 1A/5A @ 250V , Non-Inductive Operating Temperature: 20C to 50C Enclosure: IP65 / TYPE 4X / NEMA 4X

SITRANS LUT440 7MLxxxx-xxxxx-xxxx Serial No.: GYZ / B1034567 0518 KCC-REM-S49 Power Rating: 10 32V , 10W SITRANSLUT400 Contact Rating: 1A/5A @ 250V , Non-Inductive Operating Temperature: 20C to 50C Enclosure: IP65 / TYPE 4X / NEMA 4X

Nederlands

II 3D Ex tc IIIC T70C Dc Ex tc IIIC T70C Dc SIRA 12ATEX9143X IECEx SIR 12.0057X Warning: Possible static hazard. Do not rub or clean on site with a dry cloth. Warning: Do not open when energized or when an explosive dust atmosphere is present.

II 3D Ex tc IIIC T70C Dc Ex tc IIIC T70C Dc SIRA 12ATEX9143X IECEx SIR 12.0057X Warning: Possible static hazard. Do not rub or clean on site with a dry cloth. Warning: Do not open when energized or when an explosive dust atmosphere is present.

II 3D Ex tc IIIC T70C Dc Ex tc IIIC T70C Dc SIRA 12ATEX9143X IECEx SIR 12.0057X Warning: Possible static hazard. Do not rub or clean on site with a dry cloth. Warning: Do not open when energized or when an explosive dust atmosphere is present.

Siemens Milltronics Process Instruments, Peterborough Assembled in Canada with domestic and imported parts

Siemens Milltronics Process Instruments, Peterborough Assembled in Canada with domestic and imported parts

Siemens Milltronics Process Instruments, Peterborough Assembled in Canada with domestic and imported parts

Het ATEX certificaat dat is vermeld op de typeplaat kan worden gedownload van de productpagina van onze website: www.siemens.com/sitransLUT400. Het IECEx certificaat dat is vermeld op het apparaatlabel kan worden bekeken op de IECEx website. Ga naar: http://iecex.iec.ch en klik opEx Equipment Certificates of Conformity en voer vervolgens het certificaatnummer IECEx SIR 12.0057X in. Hanteer de lokale regelgeving voor de bekabelingsvereisten. Zie ook het ATEX certificaat zoals hierboven vermeld.

7ML19985XU81

SITRANS LUT400 (HART)

Pagina NL-35

2. Niet brandveroorzakende bekabeling (Canada)


SITRANS LUT420 7MLxxxx-xxxxx-xxxx Serial No.: GYZ / B1034567 Power Rating: 100 230V 15% 50/60 Hz, 36 VA (10 W) Contact Rating: 1A/5A @ 250V , Non-Inductive Operating Temperature: 20C to 50C Enclosure: IP65 / TYPE 4X / NEMA 4X SITRANS LUT430 7MLxxxx-xxxxx-xxxx Serial No.: GYZ / B1034567 Power Rating: 100 230V 15% 50/60 Hz, 36 VA (10 W) Contact Rating: 1A/5A @ 250V , Non-Inductive Operating Temperature: 20C to 50C Enclosure: IP65 / TYPE 4X / NEMA 4X SITRANS LUT440 7MLxxxx-xxxxx-xxxx Serial No.: GYZ / B1034567 Power Rating: 100 230V 15% 50/60 Hz, 36 VA (10 W) Contact Rating: 1A/5A @ 250V , Non-Inductive Operating Temperature: 20C to 50C Enclosure: IP65 / TYPE 4X / NEMA 4X

159134

Class I, Div.2, Gr. A, B, C & D T3C Class II, Div. 2, Gr. F & G Class III Per Drawing A5E03936871

159134

Class I, Div.2, Gr. A, B, C & D T3C Class II, Div. 2, Gr. F & G Class III Per Drawing A5E03936871

159134

Class I, Div.2, Gr. A, B, C & D T3C Class II, Div. 2, Gr. F & G Class III Per Drawing A5E03936871

Siemens Milltronics Process Instruments, Peterborough Assembled in Canada with domestic and imported parts

Siemens Milltronics Process Instruments, Peterborough Assembled in Canada with domestic and imported parts

Siemens Milltronics Process Instruments, Peterborough Assembled in Canada with domestic and imported parts

SITRANS LUT420 7MLxxxx-xxxxx-xxxx Serial No.: GYZ / B1034567 Power Rating: 10 32V , 10W Contact Rating: 1A/5A @ 250V , Non-Inductive Operating Temperature: 20C to 50C Enclosure: IP65 / TYPE 4X / NEMA 4X

SITRANS LUT430 7MLxxxx-xxxxx-xxxx Serial No.: GYZ / B1034567 Power Rating: 10 32V , 10W Contact Rating: 1A/5A @ 250V , Non-Inductive Operating Temperature: 20C to 50C Enclosure: IP65 / TYPE 4X / NEMA 4X

SITRANS LUT440 7MLxxxx-xxxxx-xxxx Serial No.: GYZ / B1034567 Power Rating: 10 32V , 10W Contact Rating: 1A/5A @ 250V , Non-Inductive Operating Temperature: 20C to 50C Enclosure: IP65 / TYPE 4X / NEMA 4X

159134

Class I, Div.2, Gr. A, B, C & D T3C Class II, Div. 2, Gr. F & G Class III Per DWG. A5E03936871

159134

Class I, Div.2, Gr. A, B, C & D T3C Class II, Div. 2, Gr. F & G Class III Per DWG. A5E03936871

159134

Class I, Div.2, Gr. A, B, C & D T3C Class II, Div. 2, Gr. F & G Class III Per DWG. A5E03936871

Siemens Milltronics Process Instruments, Peterborough Assembled in Canada with domestic and imported parts

Siemens Milltronics Process Instruments, Peterborough Assembled in Canada with domestic and imported parts

Siemens Milltronics Process Instruments, Peterborough Assembled in Canada with domestic and imported parts

CSA Class I, Div 2 aansluiting tekeningnr. A5E03936871 kan worden gedownload van de productpagina op onze website: www.siemens.com/sitransLUT400.

Nederlands

Instructies speciaal voor installaties in explosiegevaarlijke zones (Raadpleeg de Europese ATEX richtlijn 94/9 /EC, Annex II, 1/0/6)
De volgende instructies gelden voor apparatuur die valt onder certificaatnummer SIRA 12ATEX9143X: 1) 2) 3) 4) Raadpleeg de volledige handleiding voor gebruik en montage. De apparatuur is gecertificeerd voor gebruik als Categorie 3D apparatuur. De apparatuur mag worden gebruikt in explosiegevaarlijke zone 22. Deze apparatuur heeft een maximale oppervlaktetemperatuur van T70 C (in een +50C omgeving). Raadpleeg de van toepassing zijnde regels voor het selecteren van deze apparatuur m.b.t. stofexplosietemperaturen. 5) De apparatuur is gecertificeerd voor gebruik in een omgevingstemperatuurbereik van -20 tot +50C. 6) Eventuele kabeldoorvoeren moeten voldoen aan de vereisten van de Europese Richtlijn 94/9/EC voor Group II, Category 3D en de totale IP-beschermingsklasse van de behuizing handhaven. 7) De kabel moet worden geselecteerd overeenkomstig de geldende voorschriften en dusdanig dat deze bestand is tegen de maximale oppervlaktetemperatuur van de behuizing (T70C). 8) De apparatuur is niet getest als veiligheidstoestel (zoals omschreven in richtlijn 94/9/EC Annex II, bepaling 1.5). 9) Installatie en inspectie van deze apparatuur moet worden uitgevoerd door voldoende opgeleid personeel, conform de geldende normen (EN 60079-14 en EN 60079-17 in Europa). 10) Reparatie van deze apparatuur moet worden uitgevoerd door voldoende opgeleid personeel, conform de geldende normen (bijv. EN 60079-19 in Europa).

Pagina NL-36

SITRANS LUT400 (HART)

7ML19985XU81

11) Componenten die opgenomen moeten worden, of gebruikt moeten worden als vervangingen in de apparatuur, moeten worden gemonteerd door voldoende getraind personeel conform de documentatie van de producent. 12) Handmatige schakeling is mogelijk door bediening van de hoofdschakelaar in de gebouweninstallatie.

SPECIALE VOORWAARDEN VOOR EEN VEILIG GEBRUIK


Het achtervoegsel X achter het certificaatnummer heeft betrekking op de volgende speciale voorwaarde(n) voor veilig gebruik: Eventuele kabeldoorvoeren moeten voldoen aan de vereisten van de Europese Richtlijn 94/9/EC voor Group II, Category 3D en de totale IP-beschermingsklasse van de behuizing handhaven. De kabel moet worden geselecteerd overeenkomstig de geldende voorschriften en de isolatie moet bestand zijn tegen de maximale oppervlaktetemperatuur van de behuizing (T70C). De behuizing is niet-geleidend en kan, onder bepaalde extreme omstandigheden, een ontstekingsniveau van elektrostatische oplading genereren. De gebruiker moet waarborgen dat de apparatuur niet wordt genstalleerd op een locatie waarde deze bloot kan worden gesteld aan externe invloeden (zoals hogedrukstoom) die opbouw van elektrostatische lading op niet-geleidende oppervlakken kan veroorzaken. Bovendien mag de apparatuur alleen met een vochtige doek worden schoongemaakt. De apparatuur moet worden genstalleerd in een omgeving waar een klein risico bestaat tot mechanische schade of, indien dit niet mogelijk is, moeten worden beschermd met daarvoor geschikte maatregelen ter voorkoming van mechanische schade. Eventuele LUT400 serie regelaars (LUT420, LUT430, LUT440) die worden gebruikt bij op afstand gemonteerd deksel, moeten worden genstalleerd in een behuizing met geschikte beschermingsklasse die voldoet aan de Europese richtlijn 94/9/EC voor Group II, Category 3D en een IP65 of beter hebben.

Nederlands

7ML19985XU81

SITRANS LUT400 (HART)

Pagina NL-37

Notities

Nederlands
Pagina NL-38

SITRANS LUT400 (HART)

7ML19985XU81

Manual de incio rpido SITRANS LUT400 (HART)


Este manual descreve as caractersticas e funes essenciais do SITRANS LUT400 (HART 1)). Aconselhamos categoricamente que voc adquira a verso detalhada do manual para utilizar o mximo potencial de seu dispositivo. As instrues completas de operao do SITRANS LUT400 (HART), assim como o manual de comunicaes do SITRANS LUT400 (HART) , podem ser baixadas da pgina do produto, em nosso site, no endereo: www.siemens.com/ sitransLUT400. O manual impresso est disponvel com seu representante local Siemens. Quaisquer dvidas acerca do contedo deste manual devem ser enviadas a: Siemens AG Siemens Milltronics Process Instruments 1954 Technology Drive, P .O. Box 4225 Peterborough, Ontrio, Canad, K9J 7B1 E-mail: techpubs.smpi@siemens.com Copyright Siemens AG 2012. Todos os direitos reservados
Incentivamos os usurios a comprar os manuais vinculados autorizados ou a consultar as verses eletrnicas conforme projetadas e autorizadas pela Siemens Milltronics Process Instruments. A Siemens Milltronics Process Instruments no ser responsvel pelo contedo de reprodues parciais ou totais quer das verses limitadas como eletrnicas.

Excluso de responsabilidade
Apesar de termos verificado o contedo deste manual quanto conformidade com os instrumentos descritos, pode haver algumas variaes. Dessa forma, no podemos garantir conformidade integral. O contedo deste manual revisto com regularidade, e as correes so includas em edies posteriores. Todas as sugestes de aprimoramento so bem-vindas. Os dados tcnicos esto sujeitos a alteraes.

MILLTRONICS uma marca registrada da Siemens Milltronics Process Instruments.

O suporte est disponvel 24 horas por dia. Para saber o endereo, nmero de telefone e nmero de fax do Escritrio de Automao local da Siemens, acesse: www.siemens.com/automation/partner: Clique na aba Contact (Contato) para encontrar o grupo do produto (por exemplo, Automation Technology (Tecnologia de Automao)). Clique novamente na caixa suspensa do grupo do produto para acessar a famlia do produto. Por nvel: +Automation Technology > +Sensor Systems (Sistemas de Sensor) >+Process Instrumentation (Instrumentao de Processo) > +Level Measurement (Medio de Nvel) > +Continuous (Contnuo). Selecionar Ultrasonic (Ultrassnico). Por vazo: +Automation Technology > +Sensor Systems >+Process Instrumentation > +Flow Measurement (Medio de Vazo). Selecionar Ultrasonic. Selecione o pas seguido por regio/cidade. Selecione Technical Support (Suporte Tcnico) em Service (Servios). Para suporte tcnico on-line, acesse: www.siemens.com/automation/support-request Digite o nome do dispositivo (SITRANS LUT400) ou nmero do pedido; em seguida, clique em Search (Pesquisar) e selecione o tipo de produto apropriado. Clique em Next (Avanar). Digite uma palavra-chave que descreva seu assunto. Em seguida, navegue pela documentao relevante ou clique em Next para enviar por e-mail a descrio de seu assunto para a equipe de Suporte Tcnico da Siemens. Centro de Suporte Tcnico de Automao e Drives (A&D) da Siemens: telefone +49 (0)911 895 7222
1)

Suporte Tcnico

Portugus

HART uma marca comercial registrada da HART Communication Foundation.

7ML19985XU81

SITRANS LUT400 (HART)

Pgina PT-1

Diretrizes de segurana
As mensagens de advertncia devem ser observadas para garantir sua segurana pessoal bem como a de terceiros e para proteger o produto e o equipamento a ele associado. Essas mensagens de advertncia so acompanhadas de um esclarecimento do nvel de cautela a ser observado. O smbolo de ADVERTNCIA est relacionado com um smbolo de cuidado no produto e significa que o no cumprimento das precaues necessrias pode resultar em morte, ferimentos graves e/ou danos materiais considerveis. O smbolo de AVISO, utilizado quando no h nenhum smbolo de cuidado correspondente sobre o produto, significa que a falha em observar as precaues necessrias pode resultar em morte, ferimentos graves e/ou danos materiais considerveis. Observao: significa informao importante sobre o produto ou dessa parte do manual de operaes.

Conformidade com a FCC


Apenas para instalaes nos EUA: Normas da Federal Communications Commission (FCC) (Organismo norte-americano que regulamenta as comunicaes nos EUA)
ADVERTNCIA: As alteraes ou modificaes no aprovadas expressamente pela Siemens Milltronics podem anular a autoridade do usurio de utilizar o equipamento. Observaes: Este equipamento foi testado e avaliado em conformidade com os limites para um dispositivo digital da Classe A, de acordo com a Parte 15 das Regras da FCC. Estes limites foram concebidos para proporcionar uma proteo razovel contra interferncia nociva quando o equipamento utilizado em um ambiente comercial. Este equipamento gera, utiliza e pode irradiar energia de radiofrequncia e, se no for instalado e utilizado de acordo com o manual de instrues, pode causar interferncias nocivas nas comunicaes via rdio. provvel que o funcionamento deste equipamento em uma rea residencial cause interferncias nocivas nas comunicaes via rdio; nesse caso, o usurio precisar corrigir a interferncia sua prpria custa.

Portugus
Pgina PT-2

SITRANS LUT400 (HART)

7ML19985XU81

SITRANS LUT400
ADVERTNCIA: O SITRANS LUT400 deve ser utilizado apenas da forma descrita neste manual, caso contrrio, a proteo fornecida pelo equipamento pode ser prejudicada. Observao: Este produto foi projetado para uso em reas industriais. A operao deste equipamento em uma rea residencial pode causar interferncia em vrias comunicaes com base em frequncia. Os controladores da srie SITRANS LUT400 so controladores ultrassnicos compactos de longo alcance, com uma ponta nica para mensurao contnua de nvel de lquidos, lama e slidos e monitoramento de alta preciso de vazo de canal aberto. A srie compatvel com a toda a linha de transdutores EchoMaxda Siemens, permitindo um intervalo de alcance de 0,3 a 60 metros (dependente de transdutor). O SITRANS LUT400 foi complementado com um monitor de interface de usurio local (IUL) iluminado, com programao orientada por menu. Dispe de uma interface intuitiva com base na web, e uma gama de assistentes para desempenho plug and play. O SITRANS LUT400 compatvel com o protocolo de comunicao HART e diversos pacotes para comunicaes remotas [SIMATIC PDM, dispositivo de gerenciamento AMS, comunicador de campo 375/475 e ferramentas de dispositivo de campo (FDTs, pela sigla em ingls)]. Tambm dispe da nossa prxima gerao do Sonic Intelligence, consolidando ainda mais nossa liderana no setor em desempenho de medio, ao mesmo tempo em que aperfeioa a facilidade na aplicao.

Especificaes Alimentao

Para obter uma listagem completa, consulte o Manual de Instrues do SITRANS LUT400 (HART).

Verso CA 100-230 V CA 15%, 50 / 60 Hz, 36 VA (10W)1) Fusvel: 5 x 20 mm, baixa interrupo, 0,25A, 250V
1)

Verso CC 10-32 V CC, 10W1 Fusvel: 5 x 20 mm, baixa interrupo, 1,6A, 125V

O consumo de energia est relacionado no mximo.

Condies ambientais
Altitude 2.000 m, mx. Grau de poluio 4 Temperatura ambiente -20 a +50 C (-4 a +122 F) Categoria de instalao II Umidade relativa Compatvel com ambientes externos (apenas com revestimento IP65/Tipo 4X/NEMA 4X)

Portugus

Aprovaes
Observaes: A placa de identificao do dispositivo relaciona as aprovaes que se aplicam ao seu dispositivo. Utilize selos de condute adequados para manter a classificao de IP ou NEMA.

Geral Perigoso No-inflamvel (Canad) prova de ignio por poeira (Internacional/Europa)

CSAUS/C, CE, FM, listado UL, C-TICK CSA Classe I, Div. 2, Grupos A, B, C, D; Classe II, Div 2, Grupos F, G; Classe III ATEX II 3D Ex tc IIIC T70 C IP65 IECEx SIR 12.0057X SIRA 12ATEX9143X SITRANS LUT400 (HART) Pgina PT-3

7ML19985XU81

Instalao
Observaes: A instalao somente dever ser efetuada por pessoal qualificado e em conformidade com os regulamentos locais vigentes. Este produto suscetvel a choque eletrosttico. Siga os procedimentos adequados de aterramento. Toda a instalao eltrica de campo deve ter um isolamento adequado de, no mnimo, 250V. Voltagem perigosa presente nos terminais do transdutor durante a operao. Os terminais de entrada CC devem ser alimentados por uma fonte que proporcione isolamento eltrico entre a entrada e a sada para que sejam respeitadas as exigncias de segurana da norma IEC 61010-1. Os contatos do terminal de rel so para uso com equipamento sem partes em funcionamento acessveis e cabeamento com isolamento compatvel de, no mnimo, 250 V. A tenso de trabalho mxima permitida entre os contatos do rel adjacente deve ser de 250 V. A caixa no metlica no fornece aterramento entre as conexes de condutor. Use buchas e jumpers do tipo de aterramento.

Instrues de montagem
As instrues de montagem so diferentes para parede, tubo e trilho DIN e dispositivos de montagem em painel de monitor remoto. Siga as instrues especficas para seu dispositivo.

Observao: Ao rotear um cabo por meio de um condute, siga as instrues de roteamento de cabo em pgina 9 antes de montar o SITRANS LUT400. Montagem na parede ou painel
Todas as configuraes do SITRANS LUT400 so enviadas com uma placa traseira de montagem. O modelo de montagem de painel vem com um display IUL e uma tampa vazia.
placa traseira de montagem

Portugus

gabinete com tampa de monitor opcional

presilhas (2 locais) presilhas (2 locais)

Observao: Prendedores para montagem na parede no includos.

Pgina PT-4

SITRANS LUT400 (HART)

7ML19985XU81

placa traseira de montagem

gabinete com tampa vazia

presilhas (2 locais) presilhas (2 locais)

Observao: Prendedores para montagem na parede no includos. parte posterior do gabinete

encaixes

Montando o gabinete na parede


1) 2) 3) Marque e faa quatro orifcios na superfcie de montagem para os quatro parafusos (fornecidos pelo cliente). Aperte com uma chave de fenda. Alinhe os encaixes na parte posterior do dispositivo com a placa traseira de montagem. Pressione o nivelamento do LUT400 contra a placa traseira e deslize para baixo para fixar no lugar.

Portugus

orifcios dos parafusos de montagem na placa traseira

Observe:
Tamanho de prendedor recomendado: parafuso M8 ou 5/16" com arruela de, no mximo, 17 mm ou dimetro externo de 5/8" Montagem recomendada: monte direto na parede. Se for usada superfcie de montagem alternada, a superfcie DEVE aguentar quatro vezes o peso do dispositivo.

7ML19985XU81

SITRANS LUT400 (HART)

Pgina PT-5

Tampa montada remotamente

tampa com display remoto

junta

4 prendedores para montagem em painel fornecidos

cabo de extenso SITRANS LUT400

Montando a tampa remota


Observao: A tampa montada remota pode ser montada at 5 m do dispositivo usando dois cabos de extenso opcionais (cada um com 2,5 m de comprimento). Consulte as instrues de operao.

1)

Portugus

2)

Usando o modelo fornecido, faa o orifcio necessrio para a tampa do monitor IUL remoto. Posicione a junta dentro da tampa, alinhando com os orifcios de montagem. Alinhe a parte traseira da tampa do display remoto com o orifcio do painel. Marque e faa quatro orifcios na superfcie de montagem para os quatro parafusos (fornecidos). Fixe com uma chave de fenda e uma chave inglesa.
Observao: Torque recomendado para que a fixao dos parafusos fique bem presa: 1,1 N m 10 polegadas-libras

Observe:
Montagem recomendada: montagem no painel, at 5 m de distncia do dispositivo. Se for usada superfcie de montagem alternada, a superfcie DEVE aguentar quatro vezes o peso do dispositivo.
Observao: Prendedores includos: parafuso M5, arruela de vedao, arruela e porca de nivelamento M5. Esses prendedores so obrigatrios para manter a classificao IP65 na tampa remota.

Pgina PT-6

SITRANS LUT400 (HART)

7ML19985XU81

Montagem em tubo

placa traseira de montagem abraadeira gabinete com tampa de monitor opcional

parafuso em u

tubo Observao: Prendedores para montagem no tubo no includos.

Montagem do gabinete
1) 2) 3) Fixe a placa traseira de montagem usando os parafusos em u, abraadeiras (fornecidas pelo cliente) compatveis com o dimetro do tubo. Aperte os parafusos com uma chave inglesa. No aperte demais, pois as placas ficam tortas ou curvadas. Isso pode dificultar a fixao do clipe LUT400 placa traseira. Fixe o dispositivo na placa traseira de montagem (conforme descrito na etapa 3 de Montando o gabinete na parede na pgina 5).

Observe:
Montagem recomendada: diretamente no tubo vertical ou horizontal. Se for usada superfcie de montagem alternada, a superfcie DEVE aguentar quatro vezes o peso do dispositivo. Dimenses recomendadas para o tubo: mximo: tubo de 3", mnimo: tubo de 3/4" Tamanhos recomendados de prendedores: Parafusos em U: mximo: tamanho do tubo de 3" com rosca M8 ou 3/8" mnimo: tamanho do tubo de 3/4" com rosca M6 ou 1/4" Parafusos sextavados: M6 ou 1/4" a M8 ou 3/8" Arruela: mximo: 16 mm ou 13/16" de dimetro externo.

Portugus

7ML19985XU81

SITRANS LUT400 (HART)

Pgina PT-7

Montagem em trilho DIN


parte posterior do gabinete

trilho DIN

presilhas (2)

orficios

deslizamento do trilho DIN

Montagem do gabinete
1) 2) 3) Posicione a parte superior do gabinete em direo ao trilho DIN e posicione sobre a parte superior do trilho. Mova o gabinete para baixo, contra as presilhas de fixao na parte traseira do gabinete para a parte superior do trilho DIN. Pressione o nivelamento do dispositivo contra o trilho DIN para prender a pea deslizante do trilho DIN, o que fixar o gabinete ao trilho DIN de modo seguro.

Observe: Portugus
Montagem recomendada: diretamente na horizontal do trilho DIN. Dimenses necessrias do trilho DIN: TH 35-7.5 ou TH 35-15 por padro IEC 60715. O trilho DIN DEVE suportar at quatro vezes o peso do SITRANS LUT400.

Removendo o gabinete
1) Da parte frontal do dispositivo, posicione a chave de fenda em um orifcio na pea deslizante do trilho DIN e force para baixo. A pea deslizante soltar da parte inferior do trilho DIN. Enquanto estiver segurando a pea deslizante para baixo, puxe para cima, no sentido do gabinete, para soltar as presilhas da parte superior do trilho DIN.

2)

Pgina PT-8

SITRANS LUT400 (HART)

7ML19985XU81

Preparao para entrada do cabo


Os cabos podem ser roteados pelo condute ou coloque o gabinete por meio das gaxetas do cabo. Siga as etapas 1 a 5 abaixo para descobrir os orifcios de entrada do cabo, e depois complete as etapas para uso com condute ou gaxetas de cabo. 1) A tampa do gabinete deve estar fechada e os parafusos presos. 2) Posicione a ponta da chave de fenda na fenda do dimetro externo da guia de desativao (veja as ilustraes abaixo). 3) Bata na parte final da chave de fenda com a palma da mo para desativar o orifcio de entrada. 4) Afrouxe os parafusos e remova a tampa do gabinete. 5) Remova a(s) pea(s) de plstico que cobrem os orifcios de entrada do gabinete. Cuidado para no danificar os circuitos eletrnicos com eletricidade esttica, ou as ferramentas usadas para desativar os orifcios de entrada. Observaes:
SITRANS LUT400

orifcios de entrada do cabo

aba de desativao (3 locais)

fenda para chave de fenda (3 locais)

Quando as guias de desativao do orifcio de entrada do cabo tiverem sido removidas, o orifcio de entrada tem dimetro de 21,4 mm a 21,6 mm. As gaxetas de cabo M20 (20 mm de dimetro) e o condute 1/2" NPT (21,3 mm de dimetro) ajustam-se a esse orifcio. Cuidado ao selecionar os vedantes apropriados para os orifcios de entrada. Recomendase o uso de juntas chatas (em vez de anis em O). Se forem usadas gaxetas de cabo alternativas, de responsabilidade do cliente manter a classificao IP65 para os orifcios dos prendedores.

Cabo roteado atravs do condute


(continua das etapas 1 a 5 acima) 6) Depois de preparar para a entrada o cabo nas etapas 1 a 5, prenda o condute ao gabinete usando apenas conexes de tamanho compatveis para aplicaes prova da gua. (Tamanho do condute 1/2" NPT.) 7) Substitua a tampa do gabinete com parafusos apertados.

Exposio e insero do cabo atravs das gaxetas do cabo


(continua das etapas 1 a 5 acima) 6) Depois de preparar a entrada do cabo nas etapas 1 a 5 acima, desparafuse as gaxetas e as coloque firmemente no gabinete. 7) Rosqueie os cabos atravs das gaxetas. Verifique se o cabo de alimentao est separado dos cabos do sinal, e conecte os cabos aos blocos terminais. 8) Aperte as gaxetas de forma a fornecer uma boa vedao. 9) Substitua a tampa do gabinete com parafusos apertados.

Portugus

7ML19985XU81

SITRANS LUT400 (HART)

Pgina PT-9

Conectando
Notas de segurana para conexo
Para obter instrues detalhadas sobre a instalao eltrica, consulte o manual de instrues completo. Verifique se todos os componentes do sistema esto instalados em conformidade com as instrues. Conecte todas as blindagens do cabo aos terminais de blindagem do LUT400 (marcado ). Para evitar potenciais diferenas de atterramento, as blindagens dos pelo smbolo cabos devem estar conectadas de modo apropriado ao aterramento. Mantenha os condutores expostos nos cabos blindados assim que for possvel para reduzir o rudo na linha causado por transmisses de pulverizao e captao de rudos.

Conectando SITRANS LUT400


ADVERTNCIAS:

Examine a placa de identificao do instrumento, para verificar a classificao aprovada. Utilize selos de condute adequados para manter a classificao de IP ou NEMA.
Observaes: Podem ser necessrios cabos e condutes separados de acordo com as prticas de instalao eltrica de instrumentao padro ou cdigos de instalao eltrica. Para acessar o compartimento de instalao eltrica: 1) 2) 3) Solte dos parafusos de travamento. Levante a tampa e mova para a esquerda em suas dobradias. A tampa pode continuar aberta conectada pelas dobradias, ou pode ser desafixada das dobradias e ajustada para um lado, para o compartimento de acesso de instalao eltrica. Faa todas as conexes conforme as instrues que seguem. Quando a instalao eltrica estiver completa, substitua a tampa do dispositivo. Aperte os parafusos firmemente.

Portugus

4) 5) 6)

Pgina PT-10

SITRANS LUT400 (HART)

7ML19985XU81

Compartimento de instalao eltrica


A placa de terminal no LUT400 permite que todas as entradas e sadas sejam conectadas simultaneamente. As faixas do terminal podem ser removidas para melhorar a facilidade da instalao eltrica. CUIDADO: As faixas do terminal so terminadas localizao correta durante a reinstalao. A falha em seguir esse procedimento resulta em danos ao dispositivo, ou ao equipamento externo que est anexado. Observao: Torque recomendado dos parafusos de fixao de terminal. 0,56 0,79 N m 5 - 7 polegadas-libras No aperte os parafusos excessivamente. energia
TS-3 TRANSDUCER DISCRETE INPUTS INPUT POWER

fonte de alimentao

8 1

9 2

10

11 0.25 A, 250 V SLOW BLOW

12 L1

13 L2/N 100 - 230 V

14

DC SYNC

SHLD

WHT

BLK

BLK

WHT

bloco de terminais
mA OUT/HART BATTERY BR2032

bateria

RELAYS

DISPLAY 15 16 RLY 1 17 18 19 20 21 22 23 24

RLY 2

RLY 3 ACT

PAS SERVICE

conexo USB

cabo do monitor

conexo mA HART

Alimentao
ADVERTNCIAS Os terminais de entrada CC devem ser alimentados por uma fonte que proporcione isolamento eltrico entre a entrada e a sada para que sejam respeitadas as exigncias de segurana da norma IEC 61010-1 Toda a fiao de campo deve ter um isolamento adequado s tenses especificadas. Importante! Antes de aplicar a energia no SITRANS LUT400 pela primeira vez, assegure-se de que qualquer equipamento de alarme/controle conectado esteja desativado at que se verifique a operao e desempenho satisfatrio do sistema.

Portugus

7ML19985XU81

SITRANS LUT400 (HART)

Pgina PT-11

O SITRANS LUT400 est disponvel em modelos de energia CA ou CC.

Energia CA
INPUT POWER

Energia CC
INPUT POWER

12

13

14

12 10 - 32 V

13

14

L1 L2/N 100 - 230 V

CA: 100-230 V CA 15%, 50/60 Hz, 36 VA (10W)

CC: 10-32 V CC, 10W

Observao: Verifique se o dispositivo est conectado a um aterramento confivel. 1) 2) 3) Para fazer a instalao para energia, retire a cobertura do cabo aproximadamente 70 mm (2,75") a partir do seu extremo e passe os fios pela gaxeta 1). Ligue os fios aos terminais, conforme exibido: a polaridade est indicada abaixo do bloco de terminais. Ligue o instrumento terra de acordo com os regulamentos locais. Observaes para as conexes de energia CA: O equipamento deve ser protegido por um fusvel de 15 A ou disjuntor na instalao do edifcio. Um disjuntor ou interruptor na instalao do edifcio, identificado como interruptor de desconexo, dever estar prximo do equipamento e acessvel ao operador.

A bateria
O SITRANS LUT400 fornecido com uma bateria instalada. (Veja a ilustrao em pgina 11 para a localizao da bateria.) A bateria (BR2032) tem uma expectativa de vida til de 10 anos e afetada pela temperatura ambiente. Se o LUT400 perdeu sua energia de entrada, a bateria manter a operao do relgio de tempo real do dispositivo at a energia ser restaurada. (Veja as instrues operacionais completas para substituio da bateria.)

Portugus
1)Se o cabo estiver roteado pelo condute, utilize apenas eixos aprovados de tamanho adequado para as

aplicaes prova dgua.

Pgina PT-12

SITRANS LUT400 (HART)

7ML19985XU81

Transdutores
Advertncia: Voltagem perigosa presente nos terminais do transdutor durante a operao. Execute o cabo do transdutor em um condutor de metal aterrado, separado de outras instalaes eltricas (exceto a instalao eltrica do sensor de temperatura TS-3, se aplicvel). Observaes: No use o cabo coaxial com SITRANS LUT400 devido alta tenso transmitida da blindagem do cabo coaxial No conecte os fios da blindagem e transdutores brancos do LUT400; faa a conexo em terminais separados Desconsidere os manuais de transdutores mais antigos que recomendam essas prticas
TS-3 TRANSDUCER

SHLD

BLK

WHT

Rels
Os contatos do rel so exibidos na posio desenergizada. Todos os rels podem ser configurados como lgica negativa ou positiva (consulte o parmetro 2.8.11.Relay Logic nas instrues completas de operao).
RELAYS

BLK

WHT

Falha de energia
Os rels 2,3 esto normalmente abertos. O rel 1 pode ser conectado tanto normalmente aberto quanto fechado. Caso haja perda da entrada de energia, os rels sero revertidos aos seus estados normais.

15

16 RLY 1

17

18

19

20

21

Portugus

RLY 2

RLY 3

Classificaes do rel
um rel de forma C (NO ou NC) (rel 1), 1A a 250 V CA, no indutivo, 3A a 30 V CC dois rels de forma A (NO) (rels 2,3), 5A a 250 V CA, no indutivo, 3A a 30 V CC

7ML19985XU81

SITRANS LUT400 (HART)

Pgina PT-13

Conexo do HART sada mA (HART)


mA OUT/HART

Para conexo HART ATIVA (usando fonte de alimentao integral LUT400), conecte os terminais 22 e 23.

22

23

24

Para conexo HART PASSIVA (usando fonte de alimentao externa), conecte os terminais 23 e 24.

ACT

PAS

Para obter mais informaes, consulte os parmetros de sada mA (2.5.Current Output) na seo de referncia do parmetro nas instrues completas de operao.

Conexo em instalaes de rea perigosa


Consulte pgina 35.

Comissionando o SITRANS LUT400


Recomendamos que voc configure seu aplicativo na ordem abaixo: Execute o Assistente de Incio Rpido apropriado para sua aplicao (nvel, volume, vazo). Em seguida, configure as bombas usando o Assistente de Controle de Bomba (se aplicvel). Por ltimo, configure os alarmes, ou outros controles, totalizadores e amostradores, consultando os respectivos parmetros (consulte as instrues completas de operao). H duas formas de acessar os assistentes de incio rpido: localmente (consulte Assistentes de incio rpido via IUL na pgina 17) de uma localizao remota (consulte Outros assistentes de incio rpido: na pgina 17)

Ativando o SITRANS LUT400


Ligue o instrumento. O SITRANS LUT400 iniciado automaticamente no modo Medio. Uma tela de transio mostrando primeiro o logotipo da Siemens e, em seguida, a reviso do firmware atual do IUL exibida enquanto a primeira medida est sendo processada. A primeira vez que o dispositivo configurado, voc ser solicitado a selecionar um idioma (ingls, alemo, francs, espanhol ou chins). A data e hora corretas devem ser definidas antes da configurao do instrumento. (Para modificar, defina o parmetro 2.14.Date and Time. Consulte as Instrues de Operao para obter detalhes.)

Portugus

Pgina PT-14

SITRANS LUT400 (HART)

7ML19985XU81

O monitor de LCD Monitor em modo de medio: Funcionamento normal 1) Nvel


1 10
SITRANS LUT400 LEVEL

2
1/2

4 5

18.91
1

6
2

5 SAVING PARAMETERS

Vazo
1 10 11 12
SITRANS LUT400 FLOW HEAD
1 2 3

3
1/2
L/S M

4 5 6

500313..
60.000

DAILY TOTAL,...

999999.99

5 SAVING PARAMETERS

1 etiqueta 2 valor medido (nvel, espao, distncia, volume, vazo ou direo) 3 valor sendo exibido [Vriavel primria (PV)=1 de 2, Varivel secundria (SV)=2 de 2] 4 indicador de ativao1) para PV ou SV 5 unidades 6 o grfico de barras indica o nvel 7 a regio secundria indica os rels configurados (esquerda) e as entradas discretas (direita) 8 a rea de texto exibe mensagens de status 9 indicador de status do dispositivo 10 modo de sensor (primrio) selecionado: nvel, espao, distncia, volume, direo ou vazo 11 modo de sensor secundrio = direcionado quando o modo de sensor primrio = vazo 12 valores do totalizador: monitor alterna entre totalizador dirio e totalizador de execuo

Falha atual
0 LOE
8 a rea de texto exibe um cdigo de erro e uma mensagem de erro 9 o cone servio necessrio exibido

Monitor do modo programa Visualizao de navegao


Uma barra de menu visvel indica que menu atual/ TRANSDUCER 2.1.6 nmero nome do a lista de menus est longa demais NO TRANSDUCER do parparmetro XRS-5 metro para exibir todos os itens. XPS-10 faixa A profundidade da faixa de item na XPS-15 item de item XCT-8 atual barra de menus indica o comprimento XCT-12 XPS-30 da lista de menus: uma faixa mais barra de menus profunda indica menos itens. A posio da faixa de item indica a extenso da lista de menus e a posio aproximada do item atual na lista. Uma faixa no meio da barra de menus indica que o item atual est no meio da lista. Uma faixa mais profunda indicada poucos itens.

Portugus

1)

Pressione

ou

para ativar.

7ML19985XU81

SITRANS LUT400 (HART)

Pgina PT-15

Visualizao Parmetros
menu no qual o parmetro est localizado
SENSOR
UNITS SENSOR MODE SENSOR MODE SECONDARY

Visualizao Editar
2.1.6

nmero do parmetro

TRANSDUCER
NO TRANSDUCER XPS-10 XPS-15 XCT-8 XCT-12 XPS-30

2.1.6

XRS-5

TRANSDUCER

nome do parmetro

FREQUENCY

XRS-5

valor/seleo do parmetro

Modo PROGRAM
Utilizando os botes locais do dispositivo, pressione abra o menu de nvel 1. Percorra o menu utilizando , para entrar no modo Programa e , , .

Seleo de uma opo relacionada


a) Navegue at o parmetro desejado. b) Pressione a seta DIREITA para abrir o modo Editar. A seleo atual destacada.

c) Deslize com a barra de rolagem para uma nova seleo. d) Pressione a seta DIREITA para aceitar. O monitor LCD retorna para a visualizao de parmetros e exibe a nova seleo.

Alterao de um valor numrico


a) Navegue para o parmetro desejado, por exemplo, Ponto de Calibrao Baixo (2.2.1.). b) Quando for selecionado, o valor atual exibido. c) Pressione a seta DIREITA para abrir o modo Editar. A posio do cursor destacada. d) Use as setas ESQUERDA e DIREITA para mover o cursor para a posio do digito que voc deseja mudar. e) Como cada digito destacado (selecionado), use as setas PARA CIMA e PARA
LOW CALIB. PT. 2.2.1

ESC

+ _ _ _ _ _ _ 2. 0 M

OK

Portugus

BAIXO f)

para aumentar ou diminuir o digito. e PARA

Quando o ponto decimal for selecionado, use as setas PARA CIMA BAIXO para alterar a posio do ponto decimal.

g) Para sair sem salvar as alteraes, pressione a seta ESQUERDA

continuamente

at ESC ser destacado. Pressione a seta ESQUERDA novamente para sair sem salvar as alteraes. Caso contrrio, quando o novo valor de parmetro estiver correto, pressione a seta PARA CIMA continuamente at OK ficar em destaque.

h) Pressione a seta PARA CIMA para aceitar o novo valor. O monitor LCD retorna para a visualizao de parmetros e exibe a nova seleo. Revise para exatido.

Pgina PT-16

SITRANS LUT400 (HART)

7ML19985XU81

Assistentes de incio rpido


Para configurar o SITRANS LUT400 para aplicaes de nvel, volume (moldes de recipientes padro) ou vazo use os Assistentes de incio rpido via IUL descritos abaixo. Um assistente para aplicaes que empregam moldes de recipiente mais complexas est disponvel em SIMATIC PDM. Consultar Assistentes de Incio Rpido por meio de SIMATIC PDM na pgina 33.

Outros assistentes de incio rpido:


Outros assistentes de incio rpido usando vrios pacotes de software tambm esto disponveis 1): SIMATIC PDM (HART) AMS (HART) FC375/475 (HART) FDTs (HART) Tabelas de configurao de parmetros (para registrar dados anteriores para configurar o dispositivo via assistente de incio rpido) esto disponveis em nosso site. Visite www.siemens.com/sitransLUT400 > Support (Suporte) > Application Guides (Guias de Aplicao).

Assistentes de incio rpido via IUL


1) 2) Pressione para entrar no modo Programa. Selecione Assistentes (1.), Incio Rpido (1.1.), e o incio rpido apropriado: Nvel QS (1.1.1.), Volume QS (1.1.2.) ou Vazo QS (1.1.3.). [O assistente de vazo QS ser exibido no IUL apenas para os modelos de configurao LUT430 (Bomba e Vazo) e LUT440 (OCM).] Siga as instrues e selecione Concluir para salvar as mudanas de parmetro de Incio Rpido e voltar ao menu do Programa, pressione, ento, modo de Medio. trs vezes para voltar ao

3)

Observaes: As configuraes do Assistente de Incio Rpido so interligadas e as mudanas so aplicadas apenas depois de voc selecionar Concluir na ltima etapa. No use o assistente de incio rpido para alterar parmetros individuais: consulte a seo Paramtro das instrues completas de operao. (Faa a personalizao para a sua aplicao somente quando completar o Incio Rpido.). Para obter mais informaes sobre quaisquer parmetros do assistente, consulte as instrues completas de operao.

Portugus

1. Assistentes 1.1. Incio Rpido 1.1.1. Nvel QS


Use este assistente para configurar aplicaes de nvel simples. a) Pressione a seta DIREITA nvel 1: MENU PRINCIPAL. b) Pressione a seta DIREITA c) Pressione a seta DIREITA
1)

para ativar o modo PROGRAMA e abrir o menu de duas vezes para navegar pelo item 1.1 do menu. para abrir o nvel QS.

Consulte o manual de comunicaes para SITRANS LUT400 (HART)

7ML19985XU81

SITRANS LUT400 (HART)

Pgina PT-17

d) Em cada etapa, pressione a seta PARA para aceitar valores padro e BAIXO mover-se diretamente para o prximo item, ou a seta DIREITA para abrir o modo Editar: a seleo atual destacada. e) Acesse o item desejado e pressione a seta DIREITA para armazenar as alteraes, para pressione a seta PARA BAIXO continuar. f)

QUICK START LEVEL


TRANSDUCER XRS-5
BACK CANCEL NEXT EDIT

QUICK START LEVEL


TRANSDUCER

NO TRANSDUCER XPS-10 XPS-15 XCT-8 XCT-12

XRS-5

A qualquer momento, voc pode pressionar a seta PARA CIMA seta PARA BAIXO para cancelar o assistente. Incio do Assistente de Nvel QS

para voltar ou

Mostra o tipo de Assistente a ser executado.

Opes
Transdutor

CANCEL (CANCELAR), START (INICIAR)

Especifica o transdutor Siemens conectado ao dispositivo.

Opes
Operao

NO TRANSDUCER (SEM TRANSDUTOR), XRS-5, XPS-10, XPS-15, XCT-8, XCT-12, XPS-30, XPS-40, XLT-30, XLT-60, STH Padro: NO TRANSDUCER

Define o tipo de medio (e a sada mA correspondente) necessria para a aplicao.


ponto de referncia do sensor distncia Ponto de calibrao alto espao

Portugus

nvel Ponto de calibrao baixo

Alcance longo

Modo
NVEL ESPAO DISTNCIA

Descrio
* Altura do material Distncia at a superfcie do material

Ponto de referncia
Ponto de calibrao baixo (nvel de processo vazio) Ponto de calibrao alto (nvel pleno de processo) Sensor do ponto de referncia (continua na pgina seguinte)

Pgina PT-18

SITRANS LUT400 (HART)

7ML19985XU81

(continua da pgina anterior)

Modo

Descrio
NO selecione. Se o valor de OPERAO for exibido como OUTRO, o dispositivo est configurado como um controlador de nvel, mas foi anteriormente definido para outro modo alm de NVEL, ESPAO ou DISTNCIA. O modo de operao deve ser definido para NVEL, ESPAO ou DISTNCIA para continuar com o assistente do nvel QS.

Ponto de referncia

OUTRO

Origem de temperatura

Origem da leitura da temperatura usada para ajustar a velocidade do som..

Opes
Temperatura fixa

TRANSDUCER (TRANSDUTOR), FIXED TEMPERATURE (TEMPERATURA FIXA), EXTERNAL TS-3 (EXTERNO TS-3), AVERAGE OF SENSORS (MDIA DOS SENSORES) Padro: TRANSDUCER

Use este recurso se o dispositivo de sensibilidade de temperatura no for usado.

Valor

Intervalo: -100,0 a +150,0 C Padro: +20,0 C

Este parmetro exibido apenas se a FIXED TEMPERATURE (TEMPERATURA FIXA) selecionada para Origem de temperatura. Unidades

Unidades de medida de sensor.

Opes

M, CM, MM, FT, IN Padro: M

Observao: Para fins de exemplificao, todos os valores so considerados


como sendo em metros (m). Ponto de calibrao alto

Portugus

Distncia do Ponto de Referncia do Sensor em relao ao Ponto de Calibrao Alto: normalmente processo de nvel total.

Valor

Intervalo: 0,000 a 60,000 Padro: 0,000

Ponto de calibrao baixo

Distncia do Ponto de Referncia do Sensor em relao ao Ponto de Calibrao Baixo: normalmente nvel de processo vazio.

Valor

Intervalo: 0,000 a 60,000 Padro: 60,000

7ML19985XU81

SITRANS LUT400 (HART)

Pgina PT-19

Taxa de reao

Define a velocidade de reao do dispositivo s alteraes de medio no intervalo de alcance.


Observaes:
A taxa de reao s pode ser definida atravs do Assistente de Incio Rpido, e quaisquer mudanas feitas aos parmetros Fill Rate per Minute (2.3.1.), Empty Rate per Minute (2.3.2.) ou Damping Filter (2.3.3.) aps a concluso do assistente vo substituir a configurao da taxa de reao. Taxa de reao sempre exibe em m/minuto.
SLOW (LENTO) (0,1 M/MIN) MEDIUM (MDIO) (1,0 M/MIN) FAST (RPIDO) (10 M/MIN) Padro: SLOW (0,1 M/MIN)

Opes

Use uma configurao mais rpida do que a frequncia mxima de enchimento ou esvaziamento (a que for maior). As configuraes mais lentas tm exatido mais elevada; as configuraes mais rpidas permitem mais flutuaes rpidas de nvel. Fim do Assistente de Nvel QS

Para o QS ser bem-sucedido, todas as mudanas devem ser aplicadas.

Opes

BACK (VOLTAR), CANCEL (CANCELAR), FINISH (CONCLUIR) (Monitor volta para o menu de Incio Rpido 1.1 quando o Incio Rpido for cancelado ou concludo. Se CANCEL for selecionado, nenhuma alterao gravada no dispositivo.)

Para transferir os valores de Incio Rpido para o dispositivo e voltar ao menu do programa, pressione a seta PARA BAIXO a seta ESQUERDA (Finish (Concluir)). Depois, pressione trs vezes para voltar ao modo de medio.

1.1.2. Volume QS

Use este assistente para configurar aplicaes de volume aplicando moldes de recipiente padro. a) Pressione a seta DIREITA nvel 1. MENU PRINCIPAL. para ativar o modo PROGRAMA e abrir o menu de duas vezes para navegar pelo item 1.1 do menu. e DIREITA para abrir o Volume QS.
QUICK START VOLUME
TRANSDUCER XRS-5
BACK CANCEL NEXT EDIT

Portugus

b) Pressione a seta DIREITA c) Pressione a seta PARA BAIXO

d) Em cada etapa, pressione a seta PARA BAIXO para aceitar valores padro e mover-se diretamente para o prximo item, ou a seta DIREITA para abrir o modo Editar: a seleo atual destacada. e) Acesse o item desejado e pressione a seta DIREITA f) para armazenar as alteraes, pressione a para continuar. para seta PARA BAIXO

QUICK START VOLUME


TRANSDUCER

NO TRANSDUCER XPS-10 XPS-15 XCT-8 XCT-12

XRS-5

A qualquer momento, voc pode pressionar a seta PARA CIMA para voltar ou seta ESQUERDA cancelar o assistente.

Pgina PT-20

SITRANS LUT400 (HART)

7ML19985XU81

Incio do Assistente de Volume QS

Mostra o tipo de Assistente a ser executado.

Opes
Transdutor

CANCEL, START

Especifica o transdutor Siemens conectado ao dispositivo.

Opes

NO TRANSDUCER, XRS-5, XPS-10, XPS-15, XCT-8, XCT-12, XPS-30, XPS-40, XLT-30, XLT-60, STH Padro: NO TRANSDUCER

Origem da temperatura

Origem da leitura da temperatura usada para ajustar a velocidade do som.

Opes
Temperatura fixa

TRANSDUCER, FIXED TEMPERATURE, EXTERNAL TS-3, AVGERAGE OF SENSORS Padro: TRANSDUCER

Use este recurso se o dispositivo de sensibilidade de temperatura no for usado.

Valor

Intervalo: -100,0 a +150,0 C Padro: +20,0 C

Este parmetro exibido apenas se a FIXED TEMPERATURE selecionada para Origem de temperatura. Molde do recipiente

Define o molde do recipiente e permite que o SITRANS LUT400 calcule o volume em vez do nvel. Se NONE (NENHUM) estiver selecionado, nenhuma converso de volume executada. Selecione o molde de recipiente que combine com o recipiente ou reservatrio monitorado.
NONE (NENHUM), LINEAR (LINEAR), CYLINDER (CILINDRO), PARABOLIC BOTTOM (BASE PARABLICA), HALF SPHERE BOTTOM (BASE EM MEIA ESFERA), FLAT SLOPED BOTTOM (BASE INCLINADA PLANA), PARABOLIC ENDS (TERMINAES PARABLICAS), SPHERE (ESFERA), CONICAL BOTTOM (BASE CNICA), CURVE TABLE (TABELA EM CURVA), LINEAR TABLE (TABELA LINEAR) Padro: LINEAR

Opes

Portugus

Se as opes CURVE TABLE ou LINEAR TABLE forem selecionadas, insira os valores para nvel e ponto de interrupo de volume depois de finalizar o assistente. (Consulte os parmetros 2.6.7.Table 1-8 nas instrues de operao completas. Unidades

Unidades de medida de sensor.

Opes

M, CM, MM, FT, IN Padro: M

Observao: Para fins de exemplificao, todos os valores so considerados


como sendo em metros (m).

7ML19985XU81

SITRANS LUT400 (HART)

Pgina PT-21

Ponto de calibrao alto

Distncia do Ponto de Referncia do Sensor em relao ao Ponto de Calibrao Alto: normalmente processo de nvel total.

Valor

Intervalo: 0,000 a 60,000 Padro: 0,000

Ponto de calibrao baixo

Distncia do Ponto de Referncia do Sensor em relao ao Ponto de Calibrao Baixo: normalmente nvel de processo vazio.

Valor
Taxa de reao

Intervalo: 0,000 a 60,000 Padro: 60,000

Define a velocidade de reao do dispositivo s alteraes de medio no intervalo de alcance.

Observaes:

A taxa de reao s pode ser definida atravs do Assistente de Incio Rpido, e quaisquer mudanas feitas aos parmetros Fill Rate per Minute (2.3.1.) ou Empty Rate per Minute (2.3.2.) aps a concluso do assistente vo substituir a configurao da taxa de reao. Taxa de reao sempre exibe em m/minuto.
SLOW (0,1 M/MIN) MEDIUM (1,0 M/MIN) FAST (10 M/MIN) Padro: SLOW (0,1 M/MIN)

Opes

Use uma configurao mais rpida do que a frequncia mxima de enchimento ou esvaziamento (a que for maior). As configuraes mais lentas tm exatido mais elevada; as configuraes mais rpidas permitem mais flutuaes de nvel. Dimenso A

A altura da base do recipiente quando a base for cnica, piramidal, parablica, esfrica ou inclinada plana..

Portugus

Valor
Dimenso L

Intervalo: 0,000 a 99,999 Padro: 0,000

Comprimento da seo cilndrica de um recipente com base parablica horizontal.

Valor

Intervalo: 0,000 a 99,999 Padro: 0,000

Unidades de volume

Determina as unidades de medio de volume.

Opes

L, USGAL, IMPGAL, CUM, USER DEFINED (DEFINIDO PELO USURIO) * Padro: L

* Se a opo USER DEFINED estiver selecionada, o valor deve ser definido depois da finalizao do assistente. Consulte os parmetros 2.6.6.User Defined Unit nas instrues de operao completas. Pgina PT-22 SITRANS LUT400 (HART) 7ML19985XU81

Volume mximo

O volume mximo do recipiente. Insira o volume do recipiente correspondente ao Ponto de Calibrao Alto. Por exemplo, se o volume mximo do recipiente for 8000 L, insira o valor de 8000.

Valor

Intervalo: 0,0 a 9999999 Padro: 100,0

Fim do Assistente de Volume QS

Para o QS ser bem-sucedido, todas as mudanas devem ser aplicadas.

Opes

BACK, CANCEL, FINISH (O monitor retorna para o menu 1.1 Assistente de Incio Rpido quando o Incio Rpido for concludo ou cancelado. Se CANCEL for selecionado, nenhuma alterao gravada no dispositivo.)

Para transferir os valores de Incio Rpido para o dispositivo e voltar ao menu do programa, pressione a seta PARA BAIXO ESQUERDA (Finish). Depois, pressione a seta trs vezes para voltar ao modo de medio.

1.1.3. Vazo QS
Use este assistente para configurar aplicaes de vazo simples. [Visvel apenas nos modelos configurados LUT430 (Bomba e Vazo) e LUT440 (OCM).] a) Pressione a seta DIREITA nvel 1. MENU PRINCIPAL. b) Pressione a seta DIREITA para ativar o modo PROGRAMA e abrir o menu de duas vezes para navegar pelo item 1.1 do menu. duas vezes e a seta DIREITA para abrir a

c) Pressione a seta PARA BAIXO vazo de QS.

d) Em cada etapa, pressione a seta PARA para aceitar valores padro e BAIXO mover-se diretamente para o prximo item, ou a seta DIREITA para abrir o modo Editar: a seleo atual destacada. e) Acesse o item desejado e pressione a seta DIREITA para armazenar as alteraes, para pressione a seta PARA BAIXO continuar. f)

QUICK START FLOW


TRANSDUCER XRS-5
BACK CANCEL NEXT EDIT

Portugus

QUICK START FLOW


TRANSDUCER

NO TRANSDUCER XPS-10 XPS-15 XCT-8 XCT-12

XRS-5

A qualquer momento, voc pode pressionar a seta PARA CIMA seta PARA BAIXO para cancelar o assistente. Incio do Assistente de Vazo QS

para voltar ou

Mostra o tipo de Assistente a ser executado.

Opes

CANCEL, START

7ML19985XU81

SITRANS LUT400 (HART)

Pgina PT-23

Transdutor

Especifica o transdutor Siemens conectado ao dispositivo.

Opes

NO TRANSDUCER, XRS-5, XPS-10, XPS-15, XCT-8, XCT-12, XPS-30, XPS-40, XLT-30, XLT-60, STH Padro: NO TRANSDUCER

Origem de temperatura

Origem da leitura da temperatura usada para ajustar a velocidade do som.

Opes
Temperatura fixa

TRANSDUCER, FIXED TEMPERATURE, EXTERNAL TS-3, AVGERAGE OF SENSORS Padro: TRANSDUCER

Origem da leitura da temperatura usada para ajustar a velocidade do som.

Valor

Intervalo: -100,0 a +150,0 C Padro: +20,0 C

Este parmetro exibido apenas se a FIXED TEMPERATURE selecionada para Origem de temperatura. Dispositivo de mensurao primria

Define o dispositivo de mensurao primria (PMD) a ser usado na aplicao.


EXPONENTIAL DEVICES (DISPOSITIVOS EXPONENCIAIS), RECTANGULAR FLUME BS-3680 (CONDUTO RETANGULAR BS-3680), ROUND NOSE HORIZONTAL CR. BS-3680 (BICO HORIZONTAL ARREDONDADO CR. BS3680), TRAPEZOIDAL FLUME BS-3680 (CONDUTO TRAPEZOIDAL BS-3680), U-FLUME BS-3680 (U-CONDUTO BS-3680), FINITE CREST WEIR BS-3680 (AUDE FINITO BS-3680), THIN PLATE RECT. WEIR BS-3680 (PLACA RETANGULAR FINA. AUDE BS-3680), THIN PLATE V-NOTCH WEIR BS3680 (AUDE DE PLACA FINA COM NGULO EM V BS-3680), RECT. WEIR CONTRACTED (RETANGULAR AUDE CONTRADO), ROUND PIPE (TUBO), PALMER BOWLUS FLUME (CONDUTO PALMER BOWLUS), H-FLUME (CONDUTO-H), OTHER (OUTROS)* Padro: EXPONENTIAL DEVICES

Opes

* A opo ser definida para OTHER se o assistente foi anteriormente executado por meio da ferramenta de software HART (como SIMATIC PDM) e o dispositivo foi definido para OFF ou CABEOTE UNIVERSAL EM RELAO AO VAZO. Se esta for a configurao inicial, o PMD s pode ser definido para nenhum clculo (OFF) ou para linearizao (CABEOTE UNIVERSAL EM RELAO VAZO) via ferramentas de software HART (SIMATIC PDM, AMS, FC375/475). Consulte as instrues de operao. Mtodo de clculo de vazo

Portugus

Define o mtodo de clculo de vazo.

Opes
Unidades

ABSOLUTE (ABSOLUTO), RATIOMETRIC (RADIOMTRICO) Padro: ABSOLUTE

Unidades de medida de sensor.

Opes

M, CM, MM, FT, IN Padro: M

Observao: Para fins de exemplificao, todos os valores so considerados


como sendo em metros (m). Pgina PT-24 SITRANS LUT400 (HART) 7ML19985XU81

Ponto de calibrao alto

Distncia do Ponto de Referncia do Sensor em relao ao Ponto de Calibrao Alto: normalmente processo de nvel total.

Valor

Intervalo: 0,000 a 60,000 Padro: 0,000

Ponto de calibrao baixo

Distncia do Ponto de Referncia do Sensor em relao ao Ponto de Calibrao Baixo: normalmente nvel de processo vazio.

Valor
Taxa de reao

Intervalo: 0,000 a 60,000 Padro: 60,000

Define a velocidade de reao do dispositivo s alteraes de medio no intervalo de alcance.

Observaes:

A taxa de reao s pode ser definida atravs do Assistente de Incio Rpido, e quaisquer mudanas feitas aos parmetros Fill Rate per Minute (2.3.1.) ou Empty Rate per Minute (2.3.2.) aps a concluso do assistente vo substituir a configurao da taxa de reao. Taxa de reao sempre exibe em m/minuto.
SLOW (0,1 M/MIN) MEDIUM (1,0 M/MIN) FAST (10 M/MIN) Padro: SLOW (0,1 M/MIN)

Opes

Use uma configurao mais rpida do que a frequncia mxima de enchimento ou esvaziamento (a que for maior). As configuraes mais lentas tm exatido mais elevada; as configuraes mais rpidas permitem mais flutuaes de nvel. Fatores de clculo:

Observaes:

Os cinco parmetros abaixo sero exibidos no assistente com base no PMD selecionado acima. Esses parmetros so usados na frmula de clculo de vazo (consulte as instrues completas de operao para saber detalhes.) Expoente de vazo (PMD = DISPOSITIVOS EXPONENCIAIS)

Portugus

O expoente para frmula de clculo de vazo.

Valor

Intervalo: -999,000 a 9999,000 Padro: 1,550

7ML19985XU81

SITRANS LUT400 (HART)

Pgina PT-25

Fator K (PMD = DISPOSITIVOS EXPONENCIAIS)

A constante usada na frmula de clculo de vazo para clculo absoluto de um dispositivo exponencial, apenas.

Valor

Intervalo: -999,000 a 9999,000 Padro: 1,000

ngulo em forma de V (PMD = AUDE DE PLACA FINA COM NGULO EM V)

O ngulo em forma de V usado na frmula de clculo de vazo.

Valor
Inclinao

Intervalo: 25,000 a 95,000 Padro: 25,000

(PMD = CONDUTO TRAPEZOIDAL OU TUBO)

A inclinao de vazo usada na frmula de clculo de vazo.

Valor

Intervalo: -999,000 a 9999,000 Padro: 0,000

Coeficiente de aspereza (PMD = TUBO)

O coeficiente de aspereza de vazo usado na frmula de clculo do vazo.

Valor

Intervalo: -999,000 a 9999,000 Padro: 0,000

Portugus
Pgina PT-26

SITRANS LUT400 (HART)

7ML19985XU81

Dimenses PMD

Observaes:

Para cada PMD excluindo Dispositivos Exponenciais e Outros, voc precisa inserir at quatro dimenses. No assistente, voc ser solicitado para cada dimenso requisitada para o PMD selecionado, e o nome da dimenso do PMD respectivo ser exibido.
PMD selecionado Nome de dimenso do assistente (referncia de menu do parmetro)
APPROACH WIDTH B (LARGURA DE ACESSO B) (2.15.4.5. OCM Dimension 1) THROAT WIDTH B (LARGURA DE GARGALO B) (2.15.4.6. OCM Dimension 2) HUMP HEIGHT P (COMPRIMENTO CRISTA P) (2.15.4.7. OCM Dimension 3) THROAT LENGTH L (COMPRIMENTO DE GARGALO L) (2.15.4.8. OCM Dimension 4)

Conduto retangular BS-3680

Aude com bico horizontal arredondado BS-3680


CREST WIDTH B (LARGURA DE AUDE B) (2.15.4.5. OCM Dimension 1) CREST HEIGHT P (ALTURA DE AUDE P) (2.15.4.6. OCM Dimension 2) CREST LENGTH L (COMPRIMENTO DE AUDE L) (2.15.4.7. OCM Dimension 3)

Conduto trapezoidal BS-3680


APPROACH WIDTH B (LARGURA DE ACESSO B) (2.15.4.5. OCM Dimension 1) THROAT WIDTH B (LARGURA DE GARGALO B) (2.15.4.6. OCM Dimension 2) HUMP HEIGHT P (COMPRIMENTO CRISTA P) (2.15.4.7. OCM Dimension 3) THROAT LENGTH L (COMPRIMENTO DE GARGALO L) (2.15.4.8. OCM Dimension 4)

Conduto em U BS-3680
APPROACH DIAMETER DA (DIMETRO DE ACESSO DA) (2.15.4.5. OCM Dimension 1) THROAT DIAMETER D (DIMETRO DE GARGALO D) (2.15.4.6. OCM Dimension 2) HUMP HEIGHT P (COMPRIMENTO CRISTA P) (2.15.4.7. OCM Dimension 3) THROAT LENGTH L (COMPRIMENTO DE GARGALO L) (2.15.4.8. OCM Dimension 4)

Aude Finito BS-3680


CREST WIDTH B (LARGURA DE AUDE B) (2.15.4.5. OCM Dimension 1) CREST HEIGHT P (ALTURA DE AUDE P) (2.15.4.6. OCM Dimension 2) CREST LENGTH L (COMPRIMENTO DE AUDE L) (2.15.4.7. OCM Dimension 3)

Portugus

Aude de placa fina retangular BS-3680


APPROACH WIDTH B (LARGURA DE ACESSO B) (2.15.4.5. OCM Dimension 1) CREST WIDTH B (LARGURA DE AUDE B) (2.15.4.6. OCM Dimension 2) CREST HEIGHT P (ALTURA DE AUDE P) (2.15.4.7. OCM Dimension 3)

Aude retangular contrado


CREST WIDTH B (LARGURA DE AUDE B) (2.15.4.5. OCM Dimension 1) (continua na pgina seguinte)

7ML19985XU81

SITRANS LUT400 (HART)

Pgina PT-27

PMD selecionado (continuao)


Tubo

Nome de dimenso do assistente (referncia de menu do parmetro)

PIPE INSIDE DIAMETER D (DIMETRO INTERIOR DO TUBO D) (2.15.4.5. OCM Dimension 1)

Conduto Palmer Bowlus


MAXIMUM FLUME WIDTH HMAX (LARGURA MXIMA DE CONDUTO HMAX) (2.15.4.5. OCM Dimension 1)

Conduto-H
MAXIMUM LISTED HEAD HMAX (MXIMA DIREO LISTADA HMAX) (2.15.4.5. OCM Dimension 1)

Direo mxima

O valor de nvel mximo associado com o PMD.

Valor

Intervalo: 0,000 a 60,000 Padro: 60,000

Offset de cabeote zeto

A diferena (positiva ou negativa) entre Ponto de calibrao baixo e cabeote zero (nvel na vazo zero).

Valor

Intervalo: -60,000 a 60,000 Padro: 0,000

Unidades de taxa de vazo

As unidades de volume usadas para mostrar a vazo total.

Opes

L/S, L/MIN, CUFT/S, CUFT/D, GAL/MIN, GAL/D, IMPGAL/MIN, IMPGAL/D, CUM/H, CUM/D, USER DEFINED * Padro: L/S

* Se a opo USER DEFINED estiver selecionada, o valor deve ser definido depois da finalizao do assistente. Consulte os parmetros 2.6.6.User Defined Unit nas instrues de operao completas. Vazo mxima em 20 mA

A taxa de vazo mxima.

Portugus

Valor

Intervalo: -999 a 9999999 Padro: 100

Taxa de vazo decimal

O nmero mximo das unidades decimais a serem mostradas.

Opes

NO DIGITS (SEM DIGITOS), 1 DIGIT (1 DIGITO), 2 DIGITS (2 DIGITOS), 3 DIGITS (3 DIGITOS) Padro: NO DIGITS

Corte de vazo baixo

Elimina a atividade do totalizador para nveis de cabeote em ou abaixo do valor de corte.

Valor

Intervalo: 0,000 a 60,000 Padro: 0,000

Pgina PT-28

SITRANS LUT400 (HART)

7ML19985XU81

Fim do Assistente de Vazo QS

Para o QS ser bem-sucedido, todas as mudanas devem ser aplicadas.

Opes

BACK, CANCEL, FINISH (O monitor retorna para o menu 1.1.1 Assistente de Incio Rpido quando o Incio Rpido for concludo ou cancelado. Se CANCEL for selecionado, nenhuma alterao gravada no dispositivo.)

Para transferir os valores de Incio Rpido para o dispositivo e voltar ao menu do programa, pressione a seta PARA BAIXO ESQUERDA (Finish). Depois, pressione a seta trs vezes para voltar ao modo de medio.

Observao: Recomenda-se que um 2.15.2.Auto Zero Head seja executado depois


da concluso do assistente para assegurar a melhor exatido. Consulte as Instrues de Operao para obter detalhes.

1.2. Controle de bomba


Use este assistente para configurar as bombas, se elas sero usadas em sua aplicao. Complete o Assistente de Incio rpido aplicvel. a) Pressione a seta DIREITA nvel 1. MENU PRINCIPAL. b) Pressione a seta DIREITA c) Pressione a seta PARA BAIXO para ativar o modo PROGRAMA e abrir o menu de para navegar pelo item 1.1 do menu. e DIREITA para abrir o controle da bomba. para aceitar valores padro e

d) Em cada etapa, pressione a seta PARA BAIXO mover-se diretamente para o prximo item, ou a seta DIREITA

para abrir o modo Editar: a seleo atual destacada. para armazenar as

e) Acesse o item desejado e pressione a seta DIREITA alteraes, pressione a seta PARA BAIXO f)

para continuar. para voltar ou

A qualquer momento, voc pode pressionar a seta PARA CIMA seta PARA BAIXO para cancelar o assistente. Incio do assistente Controle da bomba

Portugus

Mostra o tipo de Assistente a ser executado.

Opes

CANCEL, START

Nmero de bombas

Selecione o nmero de bombas a serem usadas com o controle de bomba.

Opes

NONE (NENHUM), 2 Padro: NONE

Se estiver definido NONE, o controle de bomba desabilitado.

7ML19985XU81

SITRANS LUT400 (HART)

Pgina PT-29

Rel da bomba 1

Selecione o rel atribudo para a bomba 1.

Opes
Rel da bomba 2

RELAY 2 (REL 2), RELAY 3 (REL 3) Padro: RELAY 2

Apenas visualizao. Define automaticamente o rel atribudo para a bomba 2, com base no rel selecionado para a bomba 1 na etapa anterior.
Se o rel da bomba 1 = REL 2, ento a bomba do rel 2 REL 3 Opes (apenas Se o rel da bomba 1 = REL 3, ento a bomba do rel 2 = REL 2 visualizao)

Modo de controle da bomba

Define o algoritmo de controle usado para o rel.

Controlador de nvel LUT420:


ALTERNATE DUTY ASSIST, ALTERNATE DUTY BACKUP

Controlador de vazo e bomba LUT430: Opes DUTY ASSIST, SERVICE RATIO DUTY BACKUP , FIXED DUTY ASSIST, FIXED compatveis DUTY BACKUP por modelo OCM de alta exatido LUT440:
ALTERNATE DUTY ASSIST, ALTERNATE DUTY BACKUP , SERVICE RATIO DUTY ASSIST, SERVICE RATIO DUTY BACKUP , FIXED DUTY ASSIST, FIXED DUTY BACKUP Padro (todos os modelos): ALTERNATE DUTY ASSIST ALTERNATE DUTY ASSIST, ALTERNATE DUTY BACKUP , SERVICE RATIO

Coeficiente de servio da bomba 1

Selecione o uso da bomba com base no coeficiente de tempo de EXECUO em vez de com base na ltima vez usada.

Valor

Intervalo: 0 a 255 Padro: 1

Portugus

Este parmetro mostra apenas se o algoritmo de coeficiente de servio est selecionado para o modo de controle de bomba. Coeficiente de servio da bomba 2

Selecione o uso da bomba com base no coeficiente de tempo de EXECUO em vez de com base na ltima vez usada.

Valor

Intervalo: 0 a 255 Padro: 1

Este parmetro mostra apenas se o algoritmo de coeficiente de servio est selecionado para o modo de controle de bomba.

Pgina PT-30

SITRANS LUT400 (HART)

7ML19985XU81

Rel de tempo de execuo 2

Define o tempo que o rel 2 da bomba foi executado, definido em horas.

Valor

Intervalo: 0 a 999999 Padro: 0

Use o valor padro para novas bombas ou defina este valor para bombas existentes com tempo de execuo acumulado. (Este parmetro mostra apenas se o algoritmo de coeficiente de servio est selecionado para o modo de controle de bomba.) Rel de tempo de execuo 3

Define o tempo que o rel 3 da bomba foi executado, definido em horas.

Valor

Intervalo: 0 a 999999 Padro: 0

Use o valor padro para novas bombas ou defina este valor para bombas existentes com tempo de execuo acumulado. (Este parmetro mostra apenas se o algoritmo de coeficiente de servio est selecionado para o modo de controle de bomba.) Ponto de ajuste da bomba 1 ON

O nvel no qual a bomba 1 passa para ON, definido em 2.1.1.Units.

Valor

Intervalo: 0,000 a 99999,000 Padro: 0,000

Ponto de ajuste da bomba 2 ON

O nvel no qual a bomba 2 passa para ON, definido em 2.1.1.Units.

Valor

Intervalo: 0,000 a 99999,000 Padro: 0,000

Ponto de ajuste da bomba 1 OFF

O nvel no qual a bomba 1 passa para OFF, definido em 2.1.1.Units.

Valor

Intervalo: 0,000 a 99999,000 Padro: 0,000

Ponto de ajuste da bomba 2 OFF

O nvel no qual a bomba 2 passa para OFF, definido em 2.1.1.Units.

Portugus

Valor

Intervalo: 0,000 a 99999,000 Padro: 0,000

Fim do assistente Controle de bomba

Para que o assistente seja bem sucedido, todas as mudanas devem ser aplicadas.

Opes

BACK, CANCEL, FINISH (O monitor retorna para o menu de controle da bomba quando o Assistente for concludo ou cancelado. Se CANCEL for selecionado, nenhuma alterao gravada no dispositivo.)

Para transferir os valores do dispositivo e voltar ao menu do programa, pressione a seta PARA BAIXO (Finish). Depois, pressione a seta ESQUERDA para voltar ao modo de medio. duas vezes

7ML19985XU81

SITRANS LUT400 (HART)

Pgina PT-31

Comunicaes SITRANS LUT400: HART


O SITRANS LUT400 pode ser configurado em uma rede HART usando o Comunicador HART 375/475 da Emerson (consulte as instrues de operaes completas para obter detalhes) ou o pacote de software. O pacote de software recomendado o Gerenciador de Dispositivo de Processo (PDM, pela sigla em ingls) SIMATIC da Siemens. Os detalhes so descritos em um guia de aplicao Trabalhando com HART, que pode ser baixado da pgina do produto em nosso site. Acesse: www.siemens.com/sitransLUT400 em Support (Suporte) e clique em Application Guides (Guias de Aplicativos).

Operao remota
O SITRANS LUT400 compatvel com diversos softwares para operao por meio de comunicaes remotas: PC executando SIMATIC PDM PC executando o Gerenciador de Dispositivo AMS PC executando navegador PC executando uma Ferramenta de Dispositivo de Campo (FDT, pela sigla em ingls) Comunicador de campo 375/475 (FC375/FC475). Abaixo voc encontrar as informaes bsicas necessrias para usar essas ferramentas com o SITRANS LUT400. Detalhes adicionais de cada ferramenta esto disponveis no manual de comunicaes para o SITRANS LUT400 (HART).

SIMATIC PDM
O SIMATIC PDM um pacote de softwares utilizado para comissionar e preservar o SITRANS LUT400 e outros dispositivos do processo. Voc pode obter mais informaes no seguinte website: www.siemens.com/simatic-pdm. Verifique a pgina de suporte do nosso website para garantir que tem a verso mais recente do SIMATIC PDM, o Service Pack (SP) mais recente e o hot fix (HF) mais recente. Acesse: http://support.automation.siemens.com/WW/ llisapi.dll?func=cslib.csinfo&lang=en&objiD=10806857&subtype=133100

Atualizar Descrio dos Dispositivos Eletrnicos (EDD)


possvel localizar a EDD no Catlogo do Dispositivo, em Sensors/Level/Echo/Siemens AG/ SITRANS LUT400 (Sensores/Nvel/Echo/Siemens AG/SITRANS LUT400). (A EDD gravada para compatiblidade posterior.) Como uma diretriz para localizar a EDD correta, os nmeros maiores e menores devem combinar entre a reviso EDD e a reviso do firmware no dispositivo (p. ex., nmeros maiores e menores em texto em negrito: 1.00.00-02).
A reviso EDD deve corresponder reviso do Firmware no dispositivo. Para verific-la no PDM, acesse SITRANS LUT400 HART > Identification (Identificao) > Device (Dispositivo).

Portugus

combinand o firmware e revises EDD

Instalar a nova verso do SIMATIC PDM requer o Service Pack (SP) mais recente e o hot fix (HF) mais recente.

Para instalar uma nova EDD


Visite www.siemens.com/sitransLUT400 > Support (Suporte) -> Software Downloads (Downloads de Software) para baixar a EDD mais atualizada da pgina do produto do nosso site. Salve os arquivos em seu computador e extraia os arquivos comprimidos em um local de fcil acesso. Inicie o SIMATIC PDM Manager Device Catalog (Catlogo do Dispositivo Gestor), acesse e selecione a pasta que contm o arquivo EDD descompactado.

Pgina PT-32

SITRANS LUT400 (HART)

7ML19985XU81

Configurao de um novo dispositivo


1) 2)

instalar uma nova EDD, acima).

Verifique se voc tem o EDD mais recente e, se necessrio, atualize-o (consultePara Configure o SITRANS LUT400 usando o Assistente de Incio Rpido para sua aplicao.

Assistentes de Incio Rpido por meio de SIMATIC PDM


Observaes:
As configuraes do Assistente de Incio Rpido esto interligadas e as alteraes se aplicam apenas aps clicar em Apply and Transfer (Aplicar e Transferir) no fim da etapa final. Clique em BACK (VOLTAR) para voltar e revisar uma configurao ou CANCEL para sair do Incio Rpido. Abra o menu do Dispositivo em SIMATIC PDM e selecione o Incio Rpido para sua aplicao (nvel, volume, linearizao de volume, vazo).

Incio rpido

Operacional via AMS Device Manager


O AMS Device Manager um pacote de software utilizado para comissionar e preservar o SITRANS LUT400 e outros dispositivos de processo. possvel encontrar mais informaes em: http://www.emersonprocess.com/AMS/.

Descrio do Dispositivo Eletrnico (EDD)


O SITRANS LUT400 exige a EDD para o AMS Device Manager verso 10.5.

Configurar um novo dispositivo via AMS Device Manager


1) possvel localizar a EDD no Catlogo do Dispositivo, em Sensors/Level/Echo/Siemens AG/ SITRANS LUT400. Verifique a pgina do produto em nosso website: www.siemens.com/sitransLUT400, em Support-> Software Downloads, para verificar se voc tem a verso mais recente de EDD para AMS Device Manager. Salve os arquivos em seu computador e extraia os arquivos comprimidos em um local de fcil acesso. Inicie o AMS Device Manager Add Device Type, acesse o arquivo EDD descompactado e selecione-o. (Um Guia de Aplicao para a criao de dispositivos HART com AMS Device Manager pode ser baixado da pgina do produto do nosso website em Support.) Em Device Connection View (Visualizar Conexo do Dispositivo), clique com o boto direito do mouse no cone do dispositivo e selecione Scan Device (Procurar Dispositivo) para carregar os parmetros do dispositivo. Para abrir a tela inicial, clique duas vezes no cone do dispositivo. A tela inicial exibe detalhes de identificao do dispositivo e uma janela de navegao esquerda da tela. Configure o SITRANS LUT400 usando o Assistente de Incio Rpido.

Portugus

2) 3)

4)

5) 6)

7ML19985XU81

SITRANS LUT400 (HART)

Pgina PT-33

Operando via FC375/475 (HART)


O Comunicador HART FC375/FC475 um dispositivo de comunicao porttil, fcil de usar, que fornece suporte universal para outros dispositivos HART, como o SITRANS LUT400. A estrutura do menu do FC375/475 muito parecida com a do AMS Device Manager.

Descrio do Dispositivo Eletrnico (EDD)


O SITRANS LUT400 exige a EDD para FC375/FC475.

Configurar um novo dispositivo via FC375/FC475


1) Consulte a pgina do produto em nosso site em www.siemens.com/sitransLUT400, em Support->Software Downloads, to para verificar se voc tem a verso mais recente de EDD para FC375/FC475. Salve os arquivos em seu computador e extraia os arquivos comprimidos em um local de fcil acesso. Para instalar a EDD, execute o Utilitrio de Atualizao Fcil do Comunicador de Campo Emerson. Configure o SITRANS LUT400 usando o Assistente de Incio Rpido para sua aplicao.

2) 3) 4)

Operao via FDT (Ferramenta do Dispositivo de Campo)


O FDT um padro utilizado em vrios pacotes de software projetados para comissionar e manter dispositivos de campo como o SITRANS LUT400. Dois FDTs comercialmente disponveis so o PACTware e o FieldCare. O FDT muito parecido com o PDM. Para configurar um dispositivo de campo via FDT, voc precisar do DTM (Device Type Manager (Gerenciador de Tipo de Dispositivo)) para o dispositivo. Os instrumentos Siemens utilizam SITRANS DTM e um instrumento de escrita para o SITRANS DTM. 1) 2) Primeiro instale o SITRANS DTM em seu sistema. Voc pode baix-lo do website: http://www.siemens.com/sitransdtm. Clique em Support e acesse Software downloads. Instale o SITRANS LUT400 (HART) EDD para o SITRANS DTM. Pode ser baixado da pgina do produto em nosso website: Visite www.siemens.com/sitransLUT400 em Support e clique em Software Downloads.

Configurao de um novo dispositivo via FDT Portugus


Um Guia de Aplicao pode ser baixado da pgina do produto do nosso website em Support.

Servio e manuteno
O SITRANS LUT400 no necessita de manuteno ou limpeza em condies de funcionamento normais.

Reparo da unidade e excluso de responsabilidade


Para obter informaes detalhadas, consulte a contracapa interna.

Pgina PT-34

SITRANS LUT400 (HART)

7ML19985XU81

Configuraes da conexo eltrica para instalaes em reas perigosas


As opes de cabos a seguir esto disponveis para instalaes de rea de risco: Instalao eltrica prova de ignio por poeira (Internacional/Europa) Instalao eltrica no-inflamvel (Canad) Em todos os casos, verifique a placa de identificao em seu instrumento, e confirme a classificao da aprovao.

1. Instalao eltrica prova de ignio por poeira (Internacional/ Europa)


SITRANS LUT420 7MLxxxx-xxxxx-xxxx Serial No.: GYZ / B1034567 0518 KCC-REM-S49 Power Rating: 100 230V 15% SITRANSLUT400 50/60 Hz, 36 VA (10 W) Contact Rating: 1A/5A @ 250V , Non-Inductive Operating Temperature: 20C to 50C Enclosure: IP65 / TYPE 4X / NEMA 4X II 3D Ex tc IIIC T70C Dc Ex tc IIIC T70C Dc SIRA 12ATEX9143X IECEx SIR 12.0057X Warning: Possible static hazard. Do not rub or clean on site with a dry cloth. Warning: Do not open when energized or when an explosive dust atmosphere is present. Siemens Milltronics Process Instruments, Peterborough Assembled in Canada with domestic and imported parts SITRANS LUT430 7MLxxxx-xxxxx-xxxx Serial No.: GYZ / B1034567 0518 KCC-REM-S49 Power Rating: 100 230V 15% SITRANSLUT400 50/60 Hz, 36 VA (10 W) Contact Rating: 1A/5A @ 250V , Non-Inductive Operating Temperature: 20C to 50C Enclosure: IP65 / TYPE 4X / NEMA 4X II 3D Ex tc IIIC T70C Dc Ex tc IIIC T70C Dc SIRA 12ATEX9143X IECEx SIR 12.0057X Warning: Possible static hazard. Do not rub or clean on site with a dry cloth. Warning: Do not open when energized or when an explosive dust atmosphere is present. Siemens Milltronics Process Instruments, Peterborough Assembled in Canada with domestic and imported parts SITRANS LUT440 7MLxxxx-xxxxx-xxxx Serial No.: GYZ / B1034567 0518 KCC-REM-S49 Power Rating: 100 230V 15% SITRANSLUT400 50/60 Hz, 36 VA (10 W) Contact Rating: 1A/5A @ 250V , Non-Inductive Operating Temperature: 20C to 50C Enclosure: IP65 / TYPE 4X / NEMA 4X II 3D Ex tc IIIC T70C Dc Ex tc IIIC T70C Dc SIRA 12ATEX9143X IECEx SIR 12.0057X Warning: Possible static hazard. Do not rub or clean on site with a dry cloth. Warning: Do not open when energized or when an explosive dust atmosphere is present. Siemens Milltronics Process Instruments, Peterborough Assembled in Canada with domestic and imported parts

SITRANS LUT420 7MLxxxx-xxxxx-xxxx Serial No.: GYZ / B1034567 0518 KCC-REM-S49 Power Rating: 10 32V , 10W SITRANSLUT400 Contact Rating: 1A/5A @ 250V , Non-Inductive Operating Temperature: 20C to 50C Enclosure: IP65 / TYPE 4X / NEMA 4X

SITRANS LUT430 7MLxxxx-xxxxx-xxxx Serial No.: GYZ / B1034567 0518 KCC-REM-S49 Power Rating: 10 32V , 10W SITRANSLUT400 Contact Rating: 1A/5A @ 250V , Non-Inductive Operating Temperature: 20C to 50C Enclosure: IP65 / TYPE 4X / NEMA 4X

SITRANS LUT440 7MLxxxx-xxxxx-xxxx Serial No.: GYZ / B1034567 0518 KCC-REM-S49 Power Rating: 10 32V , 10W SITRANSLUT400 Contact Rating: 1A/5A @ 250V , Non-Inductive Operating Temperature: 20C to 50C Enclosure: IP65 / TYPE 4X / NEMA 4X

II 3D Ex tc IIIC T70C Dc Ex tc IIIC T70C Dc SIRA 12ATEX9143X IECEx SIR 12.0057X Warning: Possible static hazard. Do not rub or clean on site with a dry cloth. Warning: Do not open when energized or when an explosive dust atmosphere is present.

II 3D Ex tc IIIC T70C Dc Ex tc IIIC T70C Dc SIRA 12ATEX9143X IECEx SIR 12.0057X Warning: Possible static hazard. Do not rub or clean on site with a dry cloth. Warning: Do not open when energized or when an explosive dust atmosphere is present.

II 3D Ex tc IIIC T70C Dc Ex tc IIIC T70C Dc SIRA 12ATEX9143X IECEx SIR 12.0057X Warning: Possible static hazard. Do not rub or clean on site with a dry cloth. Warning: Do not open when energized or when an explosive dust atmosphere is present.

Siemens Milltronics Process Instruments, Peterborough Assembled in Canada with domestic and imported parts

Siemens Milltronics Process Instruments, Peterborough Assembled in Canada with domestic and imported parts

Siemens Milltronics Process Instruments, Peterborough Assembled in Canada with domestic and imported parts

O certificado ATEX mencionado no rtulo do dispositivo pode ser baixado na pgina de produtos em nosso site em www.siemens.com/sitransLUT400. O certificado IECEx listado na placa do dispositivo pode ser visto no website IECEx. Acesse: http://iecex.iec.ch, clique em Ex Equipment Certificates of Conformity (Certificados de Conformidade de Equipamento Ex) e introduza o nmero de certificado IECEx SIR 12.0057X.

Portugus

Para as exigncias de instalao eltrica, observe os regulamentos locais. Ver tambm o certificado ATEX listado acima.

7ML19985XU81

SITRANS LUT400 (HART)

Pgina PT-35

2. Instalao eltrica no-inflamvel (Canad)


SITRANS LUT420 7MLxxxx-xxxxx-xxxx Serial No.: GYZ / B1034567 Power Rating: 100 230V 15% 50/60 Hz, 36 VA (10 W) Contact Rating: 1A/5A @ 250V , Non-Inductive Operating Temperature: 20C to 50C Enclosure: IP65 / TYPE 4X / NEMA 4X SITRANS LUT430 7MLxxxx-xxxxx-xxxx Serial No.: GYZ / B1034567 Power Rating: 100 230V 15% 50/60 Hz, 36 VA (10 W) Contact Rating: 1A/5A @ 250V , Non-Inductive Operating Temperature: 20C to 50C Enclosure: IP65 / TYPE 4X / NEMA 4X SITRANS LUT440 7MLxxxx-xxxxx-xxxx Serial No.: GYZ / B1034567 Power Rating: 100 230V 15% 50/60 Hz, 36 VA (10 W) Contact Rating: 1A/5A @ 250V , Non-Inductive Operating Temperature: 20C to 50C Enclosure: IP65 / TYPE 4X / NEMA 4X

159134

Class I, Div.2, Gr. A, B, C & D T3C Class II, Div. 2, Gr. F & G Class III Per Drawing A5E03936871

159134

Class I, Div.2, Gr. A, B, C & D T3C Class II, Div. 2, Gr. F & G Class III Per Drawing A5E03936871

159134

Class I, Div.2, Gr. A, B, C & D T3C Class II, Div. 2, Gr. F & G Class III Per Drawing A5E03936871

Siemens Milltronics Process Instruments, Peterborough Assembled in Canada with domestic and imported parts

Siemens Milltronics Process Instruments, Peterborough Assembled in Canada with domestic and imported parts

Siemens Milltronics Process Instruments, Peterborough Assembled in Canada with domestic and imported parts

SITRANS LUT420 7MLxxxx-xxxxx-xxxx Serial No.: GYZ / B1034567 Power Rating: 10 32V , 10W Contact Rating: 1A/5A @ 250V , Non-Inductive Operating Temperature: 20C to 50C Enclosure: IP65 / TYPE 4X / NEMA 4X

SITRANS LUT430 7MLxxxx-xxxxx-xxxx Serial No.: GYZ / B1034567 Power Rating: 10 32V , 10W Contact Rating: 1A/5A @ 250V , Non-Inductive Operating Temperature: 20C to 50C Enclosure: IP65 / TYPE 4X / NEMA 4X

SITRANS LUT440 7MLxxxx-xxxxx-xxxx Serial No.: GYZ / B1034567 Power Rating: 10 32V , 10W Contact Rating: 1A/5A @ 250V , Non-Inductive Operating Temperature: 20C to 50C Enclosure: IP65 / TYPE 4X / NEMA 4X

159134

Class I, Div.2, Gr. A, B, C & D T3C Class II, Div. 2, Gr. F & G Class III Per DWG. A5E03936871

159134

Class I, Div.2, Gr. A, B, C & D T3C Class II, Div. 2, Gr. F & G Class III Per DWG. A5E03936871

159134

Class I, Div.2, Gr. A, B, C & D T3C Class II, Div. 2, Gr. F & G Class III Per DWG. A5E03936871

Siemens Milltronics Process Instruments, Peterborough Assembled in Canada with domestic and imported parts

Siemens Milltronics Process Instruments, Peterborough Assembled in Canada with domestic and imported parts

Siemens Milltronics Process Instruments, Peterborough Assembled in Canada with domestic and imported parts

CSA Classe I, Div 2, o nmero de desenho de conexo A5E03936871 pode ser baixado da pgina do produto em nosso site, no endereo www.siemens.com/sitransLUT400.

Instrues especficas para instalaes em reas perigosas (Referncia Europia ATEX Diretiva 94/9/CE, Anexo II, 1/0/6)
As instrues seguintes referem-se ao equipamento abrangido pelo certificado nmero SIRA 12ATEX9143X: 1) 2) 3) 4) Para utilizao e montagem, consulte as instrues principais. O equipamento est certificado para ser utilizado como equipamento de Categoria 3D. O equipamento pode ser utilizado em reas perigosas 22. Esse equipamento tem uma temperatura mxima na superfcie de T70 C (em uma temperatura ambiente de +50 C). Consulte o cdigo de prticas aplicvel seleo deste equipamento quanto s temperaturas especficas de ignio com poeira. 5) O equipamento est certificado para uso numa temperatura ambiente entre -20 e +50 C. 6) As entradas do cabo ou condutor devem respeitar a Diretiva Europia 94/9/CE para o Grupo II, Categoria 3D, conforme seja adequado e devero manter a classificao geral de IP para a caixa. 7) O cabo deve ser selecionado de acordo com o cdigo de prticas aplicvel de maneira que seu isolamento possa suportar o mximo de temperatura de superfcie da caixa (T70 C). 8) O equipamento no foi avaliado como dispositivo relacionado com segurana (conforme mencionado na Diretiva 94/9/CE, Anexo II, clusula 1.5). 9) A instalao e a inspeo deste equipamento devero ser efetuadas por pessoal com formao adequada, de acordo com o cdigo de prtica aplicvel (EN 60079-14 e EN 60079-17 na Europa). 10) O reparo deste equipamento dever ser efetuado por pessoal com treinamento adequado, de acordo com o cdigo de prticas aplicvel (por exemplo, EN 60079-19 na Europa). Pgina PT-36 SITRANS LUT400 (HART) 7ML19985XU81

Portugus

11) Os componentes a serem incorporados ou utilizados como componentes de substituio no equipamento devem ser ajustados por pessoal com treinamento adequado, de acordo com a documentao do fabricante. 12) A sobreposio manual pode ser realizada utilizando a chave de desconexo fornecida na instalao de fabricao.

CONDIES ESPECIAIS PARA UTILIZAO SEGURA


O sufixo X do nmero do certificado diz respeito (s) seguinte(s) condio(es) especial(is) de utilizao segura. Qualquer entrada de cabo ou condute deve atender aos requisitos da Diretriz Europeia. 94/9/CE para Grupo II, Categoria 3D e manter a classificao IP global do gabinete. O cabo deve ser selecionado de acordo com o cdigo de prticas aplicvel e seu isolamento deve suportar o mximo de temperatura de superfcie do gabinete (T70 C). O revestimento no condutivo e, sob determinadas condies extremas, pode gerar um nvel de ignio de cargas eletrostticas. O usurio deve garantir que o equipamento no seja instalado em um local onde possa estar sujeito a condies externas (como a alta presso de vapor), o que pode causar um acmulo de carga eletrosttica em superfcies no-condutoras. Alm disso, o equipamento deve ser limpo regularmente apenas com um pano mido. O equipamento deve ser instalado em uma rea onde haja baixo risco de danos mecnicos ou, se no for possvel, deve estar protegido por segurana adicional que fornea proteo mecnica adequada aplicao. Quaisquer controladores da srie LUT400 (LUT420, LUT430, LUT440) usados com tampa de montagem de painel remoto devem ser instalados em gabinetes adequadamente classificados, atendendo aos requisitos da Diretriz Europeia 94/9/CE para o Grupo II, categoria 3D e classificado como IP65 ou superior.

Portugus

7ML19985XU81

SITRANS LUT400 (HART)

Pgina PT-37

Notas

Portugus
Pgina PT-38

SITRANS LUT400 (HART)

7ML19985XU81

SITRANS LUT400 (HART) Pikakyttopas


Tm opas kertoo SITRANS LUT400 (HART 1)) -laitteen ominaisuuksista ja toiminnoista. Suosittelemme, ett hankit oppaan tydellisen version, jotta voisit kytt laitetta mahdollisimman tehokkaasti. SITRANS LUT400 (HART) -kyttoppaan laaja versio ja SITRANS LUT400 (HART) -tiedonsiirto-opas voidaan ladata tuotesivulta verkkosivuiltamme osoitteessa: www.siemens.com/sitransLUT400. Painetun oppaan saat paikalliselta Siemens-edustajaltasi. Tmn oppaan sislt koskevia asioita voi tiedustella osoitteesta: Siemens AG Siemens Milltronics Process Instruments 1954 Technology Drive, P .O. Box 4225 Peterborough, Ontario, Canada, K9J 7B1 Shkposti: techpubs.smpi@siemens.com Copyright Siemens AG 2012. Kaikki oikeudet pidtetn
Kehotamme kyttji hankkimaan oppaan hyvksytyn painetun laitoksen tai tutustumaan Siemens Milltronics Process Instrumentsin laatimaan ja hyvksymn shkiseen versioon. Siemens Milltronics Process Instruments ei vastaa osittain tai kokonaan kopioitujen painettujen tai shkisten versioiden sisllst.

Vastuuvapauslauseke
Tmn oppaan sislt on tarkistettu vastaamaan kuvattua laitteistoa, mutta muutokset ovat mahdollisia. Tmn vuoksi valmistaja ei voi taata tydellist yhtpitvyytt. Tmn oppaan sislt tarkistetaan snnllisesti ja korjaukset sisllytetn seuraaviin painoksiin. Otamme parannusehdotuksia mielellmme vastaan. Teknisiin tietoihin saattaa tulla muutoksia.

MILLTRONICS on Siemens Milltronics Process Instrumentsin rekisterity tavaramerkki.

Tekninen tuki

Tukea on saatavissa 24 tuntia vuorokaudessa. Lydt lhimmn paikallisen Siemens Automation -toimipaikan osoitteen sek puhelin- ja faksinumeron osoitteesta www.siemens.com/automation/partner: Napsauta vlilehte Contact (Yhteystiedot) ja etsi tuoteryhm (esim. Automation Technology (Automaatioteknologia)). Siirry tuoteperheeseen napsauttamalla uudelleen tuoreryhmn avattavaa luetteloruutua. Tasonmittaus: +Automation Technology > +Sensor Systems (Anturijrjestelmt) >+Process Instrumentation (Prosessi-instrumentointi) > +Level Measurement (Tasonmittaus) > +Continuous (Jatkuva). Valitse Ultrasonic (Ultrani). Virtausmittaus: +Automation Technology > +Sensor Systems >+Process Instrumentation > +Flow Measurement (Virtausmittaus). Valitse Ultrasonic. Valitse maa ja sen jlkeen kaupunki/alue (City/Region). Valitse Technical Support (Tekninen tuki) kohdasta Service (Huolto). Jos haluat teknist online-tukea, vieraile sivustossa www.siemens.com/automation/supportrequest Syt laitteen nimi (SITRANS LUT400) tai tilausnumero ja napsauta sen jlkeen Search (Etsi). Valitse sopiva tuotetyyppi. Napsauta Next (Seuraava). Syt avainsana, joka kuvaa ongelmaasi. Siirry sen jlkeen sopivaan asiakirjaan tai napsauta Next, jolloin voit lhett ongelmasi kuvauksen Siemensin tekniselle tukihenkilstlle. Siemens IA/DT Technical Support Center: puhelin +49 (0)911 895 7222
1)

Suomi

HARTon HART Communication Foundationin rekisterity tavaramerkki.

7ML19985XU81

SITRANS LUT400 (HART)

Sivu FI-1

Turvaohjeet
Annettuja varoituksia on noudatettava kyttjn ja muihin henkilihin kohdistuvien henkilvahinkojen sek tuotteen ja siihen kytkettyjen varusteiden laitevahinkojen vlttmiseksi. Varoitusten yhteydess ilmoitetaan mys vaaratasosta. VAROITUS liittyy tuotteen varoitussymboliin. Varoituksen huomiotta jttminen saattaa aiheuttaa hengenvaaran, vakavia ruumiinvammoja ja merkittvi aineellisia vahinkoja. VAROITUSTA kytetn, kun tuotteessa ei ole vastaavaa varoitussymbolia. Varoituksen huomiotta jttminen saattaa aiheuttaa hengenvaaran, vakavia ruumiinvammoja ja merkittvi aineellisia vahinkoja. Huomautus: Huomautuksissa on trkeit tuotteeseen tai kyttoppaaseen liittyvi tietoja.

FCC-sntjen mukaisuus
Ainoastaan asennuksissa Amerikan Yhdysvaltoihin: Amerikan Yhdysvaltojen liittovaltion televalvontaviranomaisen (Federal Communications Commission, FCC) snnt
VAROITUS: Laitteeseen tehdyt muutokset ja muokkaukset, joita Siemens Milltronics ei ole erikseen hyvksynyt, saattavat vaikuttaa laitteen kyttoikeuksiin. Huomautukset: Tm laite on FCC-sntjen kohdan 15 mukaisten luokan A digitaalilaitteiden vaatimusten mukainen. Nm vaatimukset on mritetty niin, ett ne suojaavat laitteen haitallisilta hiriilt kytettess laitetta kaupallisessa ympristss. Tm laite steilee ja kytt radiotaajuusenergiaa. Tm saattaa hirit radiolhetysten vastaanottoa, jos laitetta ei asenneta ja kytet kyttoppaan ohjeiden mukaan. Laitteen kyttminen asuinalueilla saattaa aiheuttaa hiriit radioliikenteeseen. Kyttj vastaa itse tllaisten hiriiden korjaamisesta.

Suomi
Sivu FI-2

SITRANS LUT400 (HART)

7ML19985XU81

SITRANS LUT400
VAROITUS: SITRANS LUT400:aa saa kytt ainoastaan tss kyttoppaassa kuvatulla tavalla, koska muutoin laitteen suojaustaso voi heikenty. Huomautus: Tm tuote on suunniteltu kytettvksi teollisuusalueilla. Laitteen kyttminen asuinalueilla saattaa aiheuttaa hiriit erilaiseen radiotaajuuksia kyttvn viestintn. Siemens SITRANS LUT400 -sarjan ohjaimet ovat pienikokoisia, yksianturisia ja mittausalueeltaan laajoja ohjaimia, joita kytetn nesteiden, lietteiden ja kiinteiden materiaalien jatkuvaan tasonmittaukseen sek avokanavavirtauksen erittin tarkkaan seurantaan. Sarja on yhteensopiva koko Siemens EchoMax -antureiden tuotevalikoiman kanssa ja mahdollistaa 0,360 metrin toiminta-alueen (anturista riippuen). SITRANS LUT400 on yhdistetty taustavalaistuun paikalliskyttliittymn (LUI) nyttn, jossa kytetn valikkopohjaista ohjelmointia. Sen erityisen etuna ovat helppokyttinen webpohjainen kyttliittym sek monet opastetut toiminnot, jotka mahdollistavat kytke ja kyt toiminnan. SITRANS LUT400 tukee HART-viestintprotokollaa ja useita etistiedonsiirron mahdollistavia ohjelmistopaketteja [SIMATIC PDM, AMS Device Manager, Field Communicator 375/475 ja FDTkenttvyltykalu (Field Device Tool)]. Siin on niin ikn kehittmmme seuraavan sukupolven Sonic Intelligence-tekniikka, joka parantaa entisestn alan huippua edustavaa mittauskyky ja helpottaa samalla kytt.

Tekniset tiedot
Virta
Vaihtovirtaversio

Tydellinen erittely lytyy SITRANS LUT400 (HART) -laitteen kyttoppaasta.

Tasavirtaversio
1032 V DC, 10 W1 Sulake: 5 x 20 mm, hidas, 1,6A, 125 V

100230 V AC 15 %, 50/60 Hz, 36 VA (10 W)1) Sulake: 5 x 20 mm, hidas, 0,25A, 250 V
1)

Tehonkulutus on ilmoitettu suurimpana arvona.

Ympristolosuhteet
Korkeus
enint. 2 000 m

Ympristn lmptila
-20+50 C (-4+122 F)

Asennusluokka
II

Likaantumisaste
4

Suhteellinen kosteus:
Soveltuu ulkokyttn (ainoastaan IP65 / Type 4X / NEMA 4X -kotelolla varustettuna)

Hyvksynnt
Huomautukset: Laitekilvess on lueteltu omalle laitteellenne mynnetyt hyvksynnt. Kyt asianmukaisia johdintiivisteit, jotta laitteen IP- tai NEMA-luokitus silyy. Yleinen Rjhdysvaaralliset tilat Sytyttmtn (Kanada) Plyss sytyttmtn (Kansainvlinen/Eurooppa) CSAUS/C, CE, FM, UL listed, C-TICK CSA luokka I, jako 2, ryhmt A, B, C, D; luokka II, jako 2, ryhmt F, G; luokka III ATEX II 3D Ex tc IIIC T70 C IP65 IECEx SIR 12.0057X SIRA 12ATEX9143X SITRANS LUT400 (HART) Sivu FI-3

Suomi

7ML19985XU81

Asennus
Huomautukset: Asennuksen saa suorittaa ainoastaan asiantunteva henkilst ja asennuksessa on noudatettava paikallisia mryksi. Tm tuote on herkk shkstaattisille purkauksille. Maadoitettava asianmukaisesti. Kaikissa kenttjohtimissa on oltava eristys ainakin 250 V:n jnnitteelle. Rjhdysvaarallisia jnnitteit esiintyy anturin liittimiss kytn aikana. Tasavirtaliittimiin on sytettv virtaa virtalhteest, jossa tulo- ja lhtpuoli on eristetty shkisesti toisistaan, jotta sovellettavat IEC 61010-1 turvallisuusvaatimukset tyttyvt. Relekoskettimien liittimi kytetn laitteistoissa, joissa ei ole paljaita jnnitteellisi osia ja joiden johdotuksen eristys on riittv vhintn 250 V:n jnnitteelle. Suurin sallittu kyttjnnite vierekkisten relekoskettimien vlill on 250 V. Epmetallinen kotelo ei toimi kaapelisuojaputkien vlisen maadoituksena. Kyt maadoitettuja lpivientiholkkeja ja kytkentjohtimia.

Asennusohjeet
Sein-, putki- ja DIN-kiskoasenteisten sek etnyttpaneelia kyttvien laitteiden asennusohjeet poikkeavat toisistaan. Noudata oman laitteesi erityisi ohjeita.

Huomautus: Jos vedt kaapelin suojaputken kautta, lue kaapelinvientiohjeet sivulla sivu 9 ennen SITRANS LUT400:n asentamista. Sein- tai paneeliasennus
Kaikkien SITRANS LUT400 -kokoonpanojen mukana toimitetaan asennustaustalevy. Paneeliasenteisen mallin mukana toimitetaan sek paikalliskyttliittymn nytt ett tavallinen kansi.
asennustaustalevy

kotelo, jossa valinnainen nyttkansi kiinnikkeet (2 paikkaa) kiinnikkeet (2 paikkaa)

Suomi

Huomautus: Seinasennuksen kiinnittimet eivt kuulu toimitukseen.

Sivu FI-4

SITRANS LUT400 (HART)

7ML19985XU81

asennustaustalevy

kotelo, jossa tavallinen kansi

kiinnikkeet (2 paikkaa) kiinnikkeet (2 paikkaa)

Huomautus: Seinasennuksen kiinnittimet eivt kuulu toimitukseen. kotelon takakansi

lovetut kohdat

Kotelon asennus seinlle


1) 2) 3) Merkitse neljn ruuvin (hankittava erikseen) paikat ja poraa asennustasoon niit varten nelj reik. Kiinnit kyttmll ruuvitalttaa. Kohdista laitteen takakannen lovetut kohdat asennustaustalevyn kiinnikkeisiin. Paina LUT400 tiiviisti taustalevy vasten ja liuuta alaspin, kunnes se kiinnittyy paikalleen.

asennusruuvien reit taustalevyss

Ota huomioon:
Suositeltava kiinnittimen koko: M8 tai 5/16" ruuvi ja aluslevy, jonka ulkohalkaisija on enintn 17 mm tai 5/8" Suositeltava asennus: asennus suoraan seinn. Jos kytt muunlaista asennusalustaa, sen TULEE kest nelj kertaa laitteen paino.

Suomi

7ML19985XU81

SITRANS LUT400 (HART)

Sivu FI-5

Etasennettava kansi

etnyttkansi

tiiviste

4 mukana toimitettua paneeliasenteista kiinnitint jatkokaapeli SITRANS LUT400

Etkannen asentaminen
Huomautus: Etasennettava kansi voidaan asentaa enintn 5 metrin phn laitteesta kyttmll kahta valinnaista jatkokaapelia (kumpikin 2,5 m pitk). Katso listiedot kyttoppaan laajasta versiosta. 1) Kyt mukana toimitettua kaavainta ja leikkaa paikallisen kyttliittymn nyttkantta varten tarvittava aukko. Aseta tiiviste kannen sisn ja kohdista asennusreit. Kohdista etnyttkannen takaosa paneeliin leikattuun reikn. Merkitse neljn ruuvin (kuuluvat toimitukseen) paikat ja poraa asennustasoon niit varten nelj reik. Kiinnit kyttmll ruuvitalttaa ja avainta. Huomautus: Kiinnitysruuvien suositeltu kiristystiukkuus hyvn lukituksen aikaansaamiseksi: 1,1 Nm 10 in-lbs

2)

Ota huomioon: Suomi


Suositeltava asennus: asenna paneeliin enintn 5 m phn laitteesta. Jos kytt muunlaista asennusalustaa, sen TULEE kest nelj kertaa laitteen paino. Huomautus: Toimitukseen kuuluvat kiinnittimet: M5-ruuvi, tiivistysaluslevy, litte M5 -aluslevy ja mutteri. Niden kiinnittimien kytt on edellytys etkannen IP65luokituksen silymiselle. Sivu FI-6 SITRANS LUT400 (HART) 7ML19985XU81

Putkiasennus

asennustaustalevy satulakiinnike

U-pultti

kotelo, jossa valinnainen nyttkansi

putki Huomautus: Putkiasennuksen kiinnittimet eivt kuulu toimitukseen.

Kotelon asentaminen
1) 2) 3) Kiinnit asennustaustalevy putkeen kyttmll putken lpimittaan sopivia U-pultteja ja satulakiinnikkeit (hankittava erikseen). Kiinnit pultit avaimella. l kirist liikaa, ettei levy vnny tai taivu. Muuten LUT400 ei kiinnity taustalevyyn. Kiinnit laite asennustaustalevyyn (kuvattu vaiheessa 3 kohdassa Kotelon asennus seinlle sivulla 5).

Ota huomioon:
Suositeltava asennus: suoraan vaaka- tai pystysuuntaiseen putkeen. Jos kytt muunlaista asennusalustaa, sen TULEE kest nelj kertaa laitteen paino. Suositeltavat putken mitat: enintn: 3" putki, vhintn: 3/4" putki Suositeltavat kiinnittimien koot: U-pultit: enintn: 3" putkikoossa M8 tai 3/8" kierre vhintn: 3/4" putkikoossa M6 tai 1/4" kierre Kuusiomutterit: M6 tai 1/4" M8 tai 3/8" Aluslevy: enintn: 16 mm tai 13/16" ulkohalkaisija.

Suomi

7ML19985XU81

SITRANS LUT400 (HART)

Sivu FI-7

DIN-kiskoasennus
kotelon takakansi

DIN-kisko

kiinnikkeet (2)

lovi

DIN-kiskon luisti

Kotelon asentaminen
1) 2) 3) Kallista kotelon ylreunaa DIN-kiskoa kohti ja aseta se juuri ja juuri kiskon ylpuolelle. Siirr koteloa alaspin DIN-kiskoa kohti, kunnes kotelon takana olevat kiinnikkeet tarttuvat DIN-kiskon ylreunaan. Paina laitetta tiiviisti DIN-kiskoa vasten, jotta DIN-kiskon luisti menee paikalleen, jolloin kotelo kiinnittyy varmasti DIN-kiskoon.

Ota huomioon:
Suositeltava asennus: suoraan vaakasuuntaiseen DIN-kiskoon. DIN-kiskon pakolliset mitat: TH 35-7.5 tai TH 35-15 standardin IEC 60715 mukaisesti. DIN-kiskon TULEE kest nelj kertaa SITRANS LUT400:n paino.

Kotelon irrottaminen
1) 2) Vie ruuvitaltta laitteen etupuolelta DIN-kiskon luistin alareunassa olevaan loveen ja vnn alaspin. Tm irrottaa luistin DIN-kiskon pohjasta. Pid luistia ala-asennossa ja tynn koteloa ylspin, jolloin kiinnikkeet irtoavat DINkiskon ylosasta.

Suomi
Sivu FI-8

SITRANS LUT400 (HART)

7ML19985XU81

Kaapelin sisnviennin valmistelu


Kaapelit voidaan vet suojaputken kautta tai vied koteloon kaapeliholkkien lpi. Tee kaapelin tuloaukkoihin reit alla lueteltujen vaiheiden 15 mukaisesti ja jatka sen mukaan, kyttk suojaputkea vai kaapeliholkkeja. 1) 2) 3) 4) 5) Varmista, ett kotelon kansi on kiinni ja ett kiinnitysruuvit on lukittu paikoilleen. Aseta ruuvitaltan krki irti pakotettavan kiekon ulkoreunalla olevaan uraan (katso alla olevasta kuvasta). Iske ruuvitaltan phn kmmenell, jolloin kiekko irtoaa lpivientiaukosta. Lys ruuveja ja irrota kotelon kansi. Irrota lpivientiaukkoja peittnyt muoviosa tai -osat kotelosta. Varo, ettei elektroniikka vahingoitu staattisesta shkst tai lpivientiaukkojen vapauttamiseen kytettvist tykaluista.
SITRANS LUT400

kaapelin lpivientiaukot

irrotettava kiekko (3 paikkaa)

ura ruuvitalttaa varten (3 paikkaa)

Huomautukset: Kun kaapelin lpivientien kiekot on poistettu, lpivientiaukko on lpimitaltaan 21,421,6 mm. Lpivientiaukkoon sopivat M20 -kaapeliholkit (lpimitta 20 mm) ja 1/2" NPT-putket (lpimitta 21,3 mm). Ole tarkka valitessasi lpivientiaukkoihin sopivia tiivisteit. Suosittelemme litte tiivistett (O-renkaan sijasta). Jos kytetn vaihtoehtoisia kaapeliholkkeja, asiakas vastaa IP65-suojausluokan silymisest lpivientiaukkojen kohdalla.

Kaapeli vedetty suojaputken kautta


(jatkoa yll oleviin vaiheisiin 15) 6) Kun olet valmistellut kaapelin lpiviennin yll olevien vaiheiden 15 mukaisesti, kiinnit suojaputki koteloon kyttmll ehdottomasti sopivan kokoisia johdintiivisteit, jotka on hyvksytty vesitiivisiin sovelluksiin. (putken koko on 1/2" NPT.) 7) Aseta kotelon kansi paikalleen ja kirist ruuvit.

Paljaan kaapelin vienti kaapeliholkkien lpi


(jatkoa yll oleviin vaiheisiin 15) 6) Kun olet valmistellut kaapelin lpiviennin yll olevien vaiheiden 15 mukaisesti, ruuvaa kaapeliholkit irti ja kiinnit ne tiukasti koteloon. 7) Pujota kaapelit holkkien lpi. Varmista, ett virtakaapeli pysyy erilln signaalikaapeleista, ja liit kaapelit riviliittimiin. 8) Kirist holkit, niin ett johdot lukittuvat tiiviisti paikoilleen. 9) Aseta kotelon kansi paikalleen ja kirist ruuvit.

Suomi

7ML19985XU81

SITRANS LUT400 (HART)

Sivu FI-9

Kytkeminen
Kytkemist koskevat turvaohjeet
Katso tydelliset kytkentohjeet kyttoppaan laajasta versiosta. Varmista, ett kaikki jrjestelmn komponentit on asennettu ohjeiden mukaisesti. Kytke kaikki kaapelivaipat LUT400:n suojaliittimiin (merkitty laitteeseen symbolilla ). Maapotentiaalien erojen vlttmiseksi on varmistettava, ett kaapelivaipat on kytketty asianmukaisesti maadoitukseen. Pid suojattujen kaapelien suojaamattomat johtimet mahdollisimman lyhyin, jotta hajasteilyn ja kohinan poiminnan aiheuttama linjakohina pysyy mahdollisimman pienen.

SITRANS LUT400:n kytkeminen


VAROITUKSET:

Tarkista laitekilvest oman mittalaitteesi hyvksyntluokitus. Kyt asianmukaisia johdintiivisteit, jotta laitteen IP- tai NEMA-luokitus silyy.
Huomautukset: Standardien tai shkasetusten mukaiseen johdotukseen saatetaan tarvita erilliset kaapelit ja johtimet. Psy kytkentkoteloon: 1) 2) 3) Lys lukitusruuveja 1/4 kierros. Nosta kansi yls ja knn se vasemmalle saranoidensa varassa. Kansi voidaan jtt auki saranoiden varaan tai se voidaan irrottaa saranoista ja asetta sivuun, jotta kytkentkoteloon pstn ksiksi.

4) 5) 6)

Tee kaikki liitnnt seuraavien ohjeiden mukaisesti. Kun johdotus on valmis, aseta laitteen kansi takaisin paikalleen. Kirist lukitusruuvit.

Suomi
Sivu FI-10

SITRANS LUT400 (HART)

7ML19985XU81

Kytkentkotelo
Kaikki lhdt ja tulot voidaan kytke LUT400:n liitinalustaan samanaikaisesti. Liitinkiskot voidaan irrottaa johdotuksen helpottamiseksi. HUOMIOITAVAA: Varmista takaisinasennuksen yhteydess, ett liitinkiskot on terminoitu oikeaan kohtaan. Laiminlynti voi johtaa laitteen tai siihen liitetyn ulkoisen varusteen vahingoittumiseen. Huomautus: Liittimien kiristysruuvien suositeltavat vntmomentit 0,560,79 Nm 57 in-lbs l kirist ruuveja liikaa. Virta
TS-3 TRANSDUCER DISCRETE INPUTS INPUT POWER

virtalhde

8 1

9 2

10

11 0.25 A, 250 V SLOW BLOW

12 L1

13 L2/N 100 - 230 V

14

DC SYNC

SHLD

WHT

BLK

BLK

WHT

riviliittimet
mA OUT/HART BATTERY BR2032

paristo

RELAYS

DISPLAY 15 16 RLY 1 17 18 19 20 21 22 23 24

RLY 2

RLY 3 ACT

PAS SERVICE

USB-liitnt

nytn kaapeli

mA HART -kytkent

Virta
VAROITUKSET: Tasavirtaliittimiin on sytettv virtaa virtalhteest, jossa tulo- ja lhtpuoli on eristetty shkisesti toisistaan, jotta sovellettavat IEC 61010-1 turvallisuusvaatimukset tyttyvt. Kaikissa kenttjohtimissa on oltava ilmoitetun jnnitteen mukainen eristys. Trke! Ennen kuin kytket virran SITRANS LUT400 -laitteeseen ensimmist kertaa varmista, ett mahdolliset hlytys- ja ohjauslaitteet ovat pois kytst. Kynnist nm vasta, kun jrjestelmn asianmukainen toimivuus on varmistettu.

Suomi

7ML19985XU81

SITRANS LUT400 (HART)

Sivu FI-11

SITRANS LUT400:sta on saatavana sek vaihtovirtamalli (AC) ett tasavirtamalli (DC).

Vaihtovirta
INPUT POWER

Tasavirta
INPUT POWER

12

13

14

12 10 - 32 V

13

14

L1 L2/N 100 - 230 V

Tasavirta: 1032 V DC, Vaihtovirta: 100230 V AC 15 %, 50/60 Hz, 10 W 36 VA (10 W) Huomautus: Varmista, ett laite on maadoitettu luotettavasti. 1) 2) 3) Virtajohto: pura kaapelin vaippaa noin 70 mm:n (2,75") matkalta toisesta pst ja pujota johdot holkin lpi. 1) Kytke johdot liittimiin yll kuvatulla tavalla: napaisuus on merkitty riviliittimen alle. Maadoita laite paikallisten vaatimusten mukaisesti. Huomautuksia vaihtovirtaliitnnist: Rakennusasennuksessa laitteisto on suojattava 15 A:n sulakkeella tai katkaisimella. Rakennusasennuksessa erottimeksi merkityn katkaisulaitteen tai kytkimen on sijaittava laitteiston lhell ja oltava helposti kyttjn ulottuvilla.

Paristo
SITRANS LUT400 toimitetaan yksi paristo asennettuna. (Katso pariston sijainti kuvasta sivu 11.) Pariston (BR2032) odotettu kyttik on kymmenen vuotta. Ympristn lmptila vaikuttaa kyttikn. Jos LUT400:n syttvirta katkeaa, paristo pit laitteen reaaliaikaisen kellon kynniss virran palautumiseen saakka. (Katso pariston vaihtamista koskevat ohjeet kyttoppaan laajasta versiosta.)

Suomi
1)Jos vedt kaapelin putken kautta, kyt vain hyvksyttyj ja sopivankokoisia vedenkestvi keskittimi.

Sivu FI-12

SITRANS LUT400 (HART)

7ML19985XU81

Anturit
Varoitus: Rjhdysvaarallinen jnnite anturin liittimiss kytn aikana. Sijoita anturin kaapeli maadoitettuun metalliputkeen ja erota tm muusta johdotuksesta (ei koske mahdollista TS-3-lmptila-anturin johdotusta). Huomautukset: l kyt SITRANS LUT400:n yhteydess koaksiaalikaapelia sen vaipan johtamien suurten jnnitteiden vuoksi. l liit LUT400:n vaippaa ja anturin valkoisia johtimia toisiinsa. Kytke ne eri liittimiin. Jt vanhojen anturioppaiden nist poikkeavat ohjeet huomiotta.
TS-3 TRANSDUCER

SHLD

BLK

WHT

Releet
Kuvan relekoskettimet ovat jnnitteettmss tilassa. Kaikki releet voidaan mritt toimimaan joko positiivisen tai negatiivisen logiikan mukaan (katso parametri 2.8.11.Relay Logic kyttoppaan laajassa versiossa).
RELAYS

BLK

WHT

Shkkatko
Releet 2, 3 ovat normaalisti auki. Rele 1 voidaan johdottaa joko normaalisti auki tai normaalisti kiinni olevaksi. Virransytn katketessa releet palaavat normaalitiloihinsa.

15

16 RLY 1

17

18

19

20

21

RLY 2

RLY 3

Releluokitukset
yksi Form C (NO tai NC) -rele (rele 1), 1A jnnitteell 250 V AC (vaihtovirta), epinduktiivinen, 3A jnnitteell 30 V DC (tasavirta) kaksi Form A (NO) -relett (releet 2,3), 5A jnnitteell 250 V AC (tasavirta), epinduktiivinen, 3A jnnitteell 30 V DC (tasavirta)

Suomi

7ML19985XU81

SITRANS LUT400 (HART)

Sivu FI-13

HART-kytkent mA-tulo (HART)


mA OUT/HART

AKTIIVISESSA HARTkytkennss (jossa kytetn LUT400:n yhdysrakenteista virtalhdett), kytke liittimiin 22 ja 23.

22

23

24

ACT

PAS

PASSIIVISESSA HART-kytkennss (jossa kytetn ulkoista virtalhdett), kytke liittimiin 23 ja 24.

Katso mA-lhtparametrien (2.5.Current Output) listiedot kyttoppaan laajan version parametritietoja ksittelevst osasta.

Kytkent rjhdysvaarallisten tilojen asennuksissa


Katso sivu 35.

SITRANS LUT400:n kyttnotto


Suosittelemme, ett sovelluksen asetukset mritetn seuraavassa jrjestyksess: Kynnist ensin sovellukseesi (taso, tilavuus, virtaus) sopiva Quick Start Wizard pikakytttoiminto. Mrit sen jlkeen tarvittaessa pumppujen asetukset pumpun ohjauksen pikakytttoiminnolla (Pump Control Wizard). Mrit lopuksi hlytykset tai muut ohjaukset, totalisaattorit ja nytteenottolaitteet vastaavien parametrien perusteella (katso tiedot kyttoppaan laajasta versiosta). Pikakytttoiminnon voi kynnist kahdella tavalla: paikallisesti (katso Ohjatut pikakytttoiminnot LUI:n avulla sivulla 17) etiskyttpaikasta (katso Muut ohjatut pikakytttoiminnot: sivulla 17)

SITRANS LUT400:n aktivoiminen


Kytke laitteeseen virta. SITRANS LUT400 kynnistyy automaattisesti mittaustilaan. Kun ensimmist mittausta ksitelln, paikalliskyttliittymn (LUI) nyttn tulee siirtymnytt, jossa nkyy aluksi Siemensin logo ja sen jlkeen nykyisen laiteohjelmiston versio. Kun laite konfiguroidaan ensimmisen kerran, sinua pyydetn valitsemaan kieli (englanti, saksa, ranska, espanja, kiina). Oikea pivmr ja kellonaika on asetettava ennen laitteen konfigurointia. (Muuta mrittmll parametri 2.14.Date and Time. Katso yksityiskohtaiset ohjeet kyttoppaan laajasta versiosta.)

Suomi
Sivu FI-14

SITRANS LUT400 (HART)

7ML19985XU81

Nestekidenytt Mittaustilan nytt: Tavallinen kytt 1) Taso


1 10
SITRANS LUT400 LEVEL

2
1/2

4 5

18.91
1

6
2

5 SAVING PARAMETERS

Virtaus
1 10 11 12
SITRANS LUT400 FLOW HEAD
1 2 3

3
1/2
L/S M

4 5 6

500313..
60.000

DAILY TOTAL,...

999999.99

5 SAVING PARAMETERS

1 tunniste 2 mittausarvo (level (taso), space (vli), distance (etisyys), volume (tilavuus), flow (virtaus) tai head (energiakorkeus)) 3 nyttarvo [ensisijainen muuttuja (PV)=1/2, toissijainen muuttuja (SV)=2/2] 4 vaihtokytkin 1) PV vai SV 5 yksikk 6 pinnankorkeuden tason palkkikuvaaja 7 toissijainen alue nytt konfiguroidut releet (vasemmalla) ja diskreetit sisntulot (oikealla) 8 tekstialue nytt tilaviestit 9 laitteen tilailmaisin 10 valittu (ensisijainen) mittapn tila: level, space, distance, volume, head tai flow 11 toissijainen mittapn tila = head, kun ensisijainen mittapn tila = flow 12 totalisaattorin arvot: nytt vaihdellen pivn totalisaattorin ja koko kytn totalisaattorin

Vika ilmennyt
0 LOE

8 tekstialue nytt virhekoodin ja virheviestin 9 huoltotarvekuvake tulee nkyviin

Program-ohjelmointitilan nytt Navigaationytt


Nkyviin tuleva valikkopalkki TRANSDUCER 2.1.6 tarkoittaa, ett valikkoluettelo on niin valikon/ NO TRANSDUCER parametrin parametrin XRS-5 nimi pitk, etteivt kaikki kohdat ny. numero XPS-10 kohdan Valikon kohtaa ilmaisevan nelin XPS-15 valittu neli XCT-8 kohta pituus osoittaa valikkoluettelon XCT-12 XPS-30 pituuden: pitempi neli tarkoittaa, ett valikko- palkki kohtia on vhemmn. Nelin sijainti osoittaa nykyisen kohdan viitteellisen sijainnin luettelossa. Kun neli on puolivliss valikkopalkkia, se tarkoittaa, ett nykyinen kohta on puolivliss luetteloa. Pitempi neli tarkoittaa, ett kohtia on vhemmn.

Suomi

1)Kytke painamalla

tai

7ML19985XU81

SITRANS LUT400 (HART)

Sivu FI-15

Parametrinytt
parametrin sijaintivalikko
SENSOR
UNITS SENSOR MODE SENSOR MODE SECONDARY

Muokkausnytt
2.1.6

parametrin numero parametrin arvo/valinta

TRANSDUCER
NO TRANSDUCER XPS-10 XPS-15 XCT-8 XCT-12 XPS-30

2.1.6

XRS-5

TRANSDUCER

parametrin nimi

FREQUENCY

XRS-5

PROGRAM-ohjelmointitila
Laitteen omia painikkeita kytten: siirry Program-ohjelmointitilaan painamalla aukeaa valikkotaso 1. Selaa valikkoa kyttmll , , , . , jolloin

Luetellun vaihtoehdon valinta


a) Siirry haluttuun parametriin. b) Paina nuoli OIKEALLE nkyy korostettuna. , niin siirryt muokkaustilaan. Valittuna oleva vaihtoehto

c) Vierit uuden valinnan kohdalle. d) Hyvksy se painamalla nuoli OIKEALLE . Nestekidenytt palaa parametrinyttn ja nytt uuden valinnan.

Numeroarvon muuttaminen
a) Siirry haluttuun parametriin, esim. Low Calibration Point (Matala kalibrointipiste) (2.2.1.). b) Valittu nykyinen arvo nkyy nytss. c) Paina nuoli OIKEALLE , niin siirryt muokkaustilaan: Kohdistimen sijainti nkyy korostettuna. d) Siirr kohdistin muutettavan numeron kohdalla painamalla nuoli VASEMMALLE painamalla joko nuoli YLS f) tai OIKEALLE tai ALAS . tai .
LOW CALIB. PT. 2.2.1

ESC

+ _ _ _ _ _ _ 2. 0 M

OK

e) Kun yksittinen numero on korostettuna (valittuna), suurenna tai pienenn numeroa

Kun desimaalipiste on valittuna, voit siirt sen sijainti painamalla nuoli YLS ALAS .

g) Voit poistua tallentamatta muutoksia painamalla nuoli VASEMMALLE toistuvasti, kunnes ESC nkyy korostettuna. Nyt voit poistua tallentamatta muutoksia painamalla nuoli VASEMMALLE uudelleen. Jos uusi parametri sit vastoin on toistuvasti, kunnes OK nkyy korostettuna. oikein, paina nuoli OIKEALLE

Suomi

h) Hyvksy uusi arvo painamalla nuoli OIKEALLE . Nestekidenytt palaa parametrinyttn ja nytt uuden valinnan. Tarkista arvon oikeellisuus. Sivu FI-16 SITRANS LUT400 (HART) 7ML19985XU81

Ohjatut pikakytttoiminnot
Mrit SITRANS LUT400 tason, tilavuuden (vakiomuotoiset asiat) tai virtauksen mittaamisen sovelluksiin kyttmll alla kuvattua menettely Ohjatut pikakytttoiminnot LUI:n avulla. Muodoiltaan monimutkaisempia astioita kyttviin sovelluksiin soveltuva ohjattu toiminto on kytettviss SIMATIC PDM -tykalun avulla. Katso Ohjatut pikakytttoiminnot SIMATIC PDM:n avulla sivulla 33.

Muut ohjatut pikakytttoiminnot:


Mys erilaisiin ohjelmistopaketteihin perustuva muita ohjattuja pikakytttoimintoja on saatavissa 1): SIMATIC PDM (HART) AMS (HART) FC375/475 (HART) FDT:t (HART) Verkkosivustossamme on saatavana parametrien mrityskaavioita (Parameter Configuration Charts) tietojen kirjaamiseksi ennen laitteen konfigurointia ohjatun pikakytttoiminnon avulla. Siirry sivulle www.siemens.com/sitransLUT400 > Support (Tuki) > Application Guides (Sovellusoppaat).

Ohjatut pikakytttoiminnot LUI:n avulla


1) 2) Siirry Program-ohjelmointitilaan painamalla . Valitse Ohjatut toiminnot (1.), Pikakytttoiminto (1.1.) ja sen jlkeen sopiva pikakytttoiminto: QS Level (Tasonmittauksen ohjattu pikakytttoiminto) (1.1.1.), QS Volume (Tilavuusmittauksen ohjattu pikakytttoiminto) (1.1.2.) tai QS Flow (Virtausmittauksen ohjattu pikakynnistystoiminto) (1.1.3.). [Virtausmittauksen ohjattu pikakytttoiminto nkyy ainoastaan LUT430:n (pumppu ja virtaus) ja LUT440:n (OCM, avokanavirtaus) vastaavasti konfiguroitujen mallien paikalliskyttliittymiss.] Noudata ohjeita ja valitse sen jlkeen Finish (Lopeta), jolloin ohjatun pikakytttoiminnon parametrin tallentuvat ja laite palaa Program-ohjelmointivalikkoon. Palaa sen jlkeen mittaustilaan painamalla kolme kertaa .

3)

Huomautukset: Ohjatun pikakytttoiminnon asetukset ovat sidoksissa toisiinsa, ja muutokset tulevat voimaan vasta, kun valitset toiminnon lopuksi Finish. l yrit muuttaa yksittisi parametreja ohjatulla pikakytttoiminnolla: katso ohjeet sen sijaan kyttoppaan laajan version parametreja koskevasta osasta. (Mukauta laite omaa sovellustasi varten vasta pikakytttoiminnon pttymisen jlkeen.) Jos tarvitset listietoja ohjatun pikakytttoiminnon yksittisist parametreista, katso kyttoppaan laajasta versiosta.

1. Ohjatut toiminnot 1.1. Pikakytttoiminto 1.1.1. QS Level (Tasonmittauksen ohjattu pikakytttoiminto)

Tll ohjatulla pikakytttoiminnolla mritetn yksinkertaisia tasonmittaussovelluksia. a) Nuoli OIKEALLE aktivoi PROGRAM-ohjelmointitilan ja avaa valikkotason 1. MAIN MENU (PVALIKKO). b) Siirry valikon kohtaan 1.1.1 painamalla kaksi kertaa nuoli OIKEALLE . c) Avaa QS Level (Tasonmittauksen ohjattu pikakytttoiminto) painamalla nuoli OIKEALLE .

Suomi

1)Katso listietoja SITRANS LUT400 (HART) -tiedonsiirto-oppaasta.

7ML19985XU81

SITRANS LUT400 (HART)

Sivu FI-17

d) Joka kohdassa voit hyvksy oletusarvot ja siirty seuraavaan kohtaan painamalla nuoli tai siirty muokkaustilaan ALAS painamalla nuoli OIKEALLE : valittuna oleva vaihtoehto nkyy korostettuna. e) Vierit halumaasi kohtaan ja tallenna muutos painamalla nuoli OIKEALLE ja jatka sen jlkeen painamalla nuoli ALAS . f)

QUICK START LEVEL


TRANSDUCER XRS-5
BACK CANCEL NEXT EDIT

QUICK START LEVEL


TRANSDUCER

NO TRANSDUCER XPS-10 XPS-15 XCT-8 XCT-12

XRS-5

Voit milloin tahansa palata taaksepin tai peruuttaa painamalla nuoli YLS ohjatun toiminnon painamalla nuoli VASEMMALLE . Start of QS Level Wizard (Tasonmittauksen ohjatun pikakytttoiminnon kynnistys)

Nytt suoritettavan ohjatun toiminnon tyypin.

Vaihtoehdot

CANCEL (PERUUTA), START (KYNNIST)

Transducer (Anturi)

Mritt laitteeseen kytketyn Siemens-anturin.

Vaihtoehdot

NO TRANSDUCER (EI ANTURIA), XRS-5, XPS-10, XPS-15, XCT-8, XCT-12, XPS-30, XPS-40, XLT-30, XLT-60, STH Oletusarvo: NO TRANSDUCER

Operation (Toiminta)

Mritt sovelluksessa tarvittavan mittaustyypin (ja vastaavan mA-lhdn).


mittapn viitepiste Korkea kalibrointipiste vli

etisyys

taso Matala kalibrointipiste

Kaukomittausalue

Tila
LEVEL (TASO) SPACE (VLI) DISTANCE (ETISYYS) *

Kuvaus
Materiaalin korkeus Etisyys materiaalin pintaan

Viitepiste
Matala kalibrointipiste (prosessin perustaso). Korkea kalibrointipiste (prosessin ylin taso) Mittapn viitepiste (jatkuu seuraavalla sivulla)

Suomi
Sivu FI-18

SITRANS LUT400 (HART)

7ML19985XU81

(jatkoa edelliselt sivulta)

Tila

Kuvaus
L valitse. Jos Operation (Kytt) arvona nkyy OTHER, laite on mritetty mittaamaan tasoa mutta on aiemmin mritetty muuhun mittaustilaan kuin LEVEL, SPACE tai DISTANCE. Toimintatavaksi on mritettv LEVEL, SPACE tai DISTANCE, jotta tasonmittauksen ohjattua pikakytttoimintoa voidaan jatkaa.

Viitepiste

OTHER (MUU)

Temperature Source (Lmptilan lhde)

nen nopeuden stmisess kytettv lmptilalukeman lhde.

Vaihtoehdot

TRANSDUCER (ANTURI), FIXED TEMPERATURE (KIINTE LMPTILA), EXTERNAL TS-3 (ULKOINEN TS-3), AVERAGE OF SENSORS (MITTAPIDEN KESKIARVO) Oletusarvo: TRANSDUCER

Fixed Temperature (Kiinte lmptila)

Kyt tt vaihtoehtoa, jollei lmptilaa mittaavaa laitetta ole kytss.

Arvo

Alue: -100,0+150,0 C Oletusarvo: +20,0 C

Tm parametri nkyy vain, jos FIXED TEMPERATURE on valittu kohdassa Temperature Source (Lmptilan lhde). Units (Yksikt)

Mittapn mittausyksikt..

Vaihtoehdot

M, CM, MM, FT, IN Oletusarvo: M

Huomautus: Tss esimerkiss kaikki arvot ilmoitetaan oletusarvoisesti


metrein (m). High Calibration Point (Korkea kalibrointipiste)

Etisyys mittapn viitepisteest korkeaan kalibrointipisteeseen: yleens prosessin ylin taso..

Arvo

Alue: 0,00060,000 Oletusarvo: 0,000

Low Calibration Point (Matala kalibrointipiste)

Suomi

Etisyys mittapn viitepisteest matalaan kalibrointipisteeseen: yleens prosessin perustaso..

Arvo
7ML19985XU81

Alue: 0,00060,000 Oletusarvo: 60,000

SITRANS LUT400 (HART)

Sivu FI-19

Response Rate (Vastenopeus)

Mritt laitteen reagointinopeuden mittaustulosten muutoksiin valitulla mittausalueella.

Huomautukset:

Vastenopeus voidaan mritt ainoastaan ohjatun pikakytttoiminnon avulla. Aina kun parametreja Fill Rate per Minute (2.3.1.), Empty Rate per Minute (2.3.2.) tai Damping Filter (2.3.3.) muutetaan ohjatun pikakytttoiminnan pttmisen jlkeen, vastenopeusasetuksen arvot korvautuvat uusilla. Response Rate (Vastenopeus) nytetn aina metrein minuutissa (m/min).
SLOW (HIDAS) (0.1 M/MIN) MEDIUM (KESKITASO) (1.0 M/MIN) FAST (NOPEA) (10 M/MIN) Oletusarvo: SLOW (0.1 M/MIN)

Vaihtoehdot

Kyt asetusta, joka on hieman nopeampi kuin tytn tai tyhjentymisen enimmisnopeus (sen mukaan, kumpi on suurempi). Hitaampi asetus on tarkempi, nopeampi asetus sallii nopeammat tasonvaihtelut End of QS Level Wizard (Tasonmittauksen ohjatun pikakytttoiminnon loppu)

Jotta ohjattu pikakytttoiminto onnistuu, kaikki muutokset on otettava kyttn.

Vaihtoehdot

BACK (TAKAISIN), CANCEL (PERUUTA), FINISH (LOPETA) (Nyttn palautuu 1.1 Quick Start -pikakyttvalikko, kun ohjattu pikakytttoiminto on Quick Start is suoritettu onnistuneesti tai peruutettu. Jos valitaan CANCEL, muutoksia ei tallenneta laitteeseen.)

Siirr pikakynnistysarvot laitteeseen ja palaa ohjelmointivalikkoon painamalla nuoli ALAS (Finish). Palaa sen jlkeen mittaustilaan painamalla kolme kertaa . nuoli VASEMMALLE

1.1.2. QS Volume (Tilavuusmittauksen ohjattu pikakytttoiminto)


a) Nuoli OIKEALLE MENU.

Tll ohjatulla pikakytttoiminnolla mritetn vakiomuotoisia astioita kyttvi tilavuusmittaussovelluksia. aktivoi PROGRAM-ohjelmointitilan ja avaa valikkotason 1. MAIN

b) Siirry valikon kohtaan 1.1.1 painamalla kaksi kertaa nuoli OIKEALLE . c) Avaa QS Volume (Tilavuusmittauksen ohjattu pikakytttoiminto) painamalla nuoli ALAS ja nuoli OIKEALLE . d) Joka kohdassa voit hyvksy oletusarvot ja siirty seuraavaan kohtaan painamalla nuoli ALAS tai siirty muokkaustilaan painamalla nuoli

QUICK START VOLUME


TRANSDUCER XRS-5
BACK CANCEL NEXT EDIT

OIKEALLE : valittuna oleva vaihtoehto nkyy korostettuna. e) Vierit halumaasi kohtaan ja tallenna muutos painamalla nuoli OIKEALLE ja jatka sen jlkeen painamalla nuoli ALAS . Voit milloin tahansa palata taaksepin painamalla nuoli YLS tai peruuttaa ohjatun toiminnon painamalla . SITRANS LUT400 (HART) nuoli VASEMMALLE Sivu FI-20

QUICK START VOLUME


TRANSDUCER

NO TRANSDUCER XPS-10 XPS-15 XCT-8 XCT-12

XRS-5

Suomi

f)

7ML19985XU81

Start of QS Volume Wizard (Tilavuusmittauksen ohjatun pikakytttoiminnon kynnistys)

Nytt suoritettavan ohjatun toiminnon tyypin.

Vaihtoehdot CANCEL, START


Transducer (Anturi)

Mritt laitteeseen kytketyn Siemens-anturin.


NO TRANSDUCER, XRS-5, XPS-10, XPS-15, XCT-8, XCT-12, XPS-30, XPS-40,

Vaihtoehdot XLT-30, XLT-60, STH


Oletusarvo: NO TRANSDUCER

Temperature Source (Lmptilan lhde)

nen nopeuden stmisess kytettv lmptilalukeman lhde


TRANSDUCER, FIXED TEMPERATURE, EXTERNAL TS-3, AVERAGE OF

Vaihtoehdot SENSORS
Oletusarvo: TRANSDUCER

Fixed Temperature (Kiinte lmptila)

Kyt tt vaihtoehtoa, jollei lmptilaa mittaavaa laitetta ole kytss.

Arvo

Alue: -100,0+150,0 C Oletusarvo: +20,0 C

Tm parametri nkyy vain, jos FIXED TEMPERATURE on valittu kohdassa Temperature Source (Lmptilan lhde). Vessel Shape (Astian muoto)

Mritt astian muodon, jolloin SITRANS LUT400 pystyy laskemaan tason sijasta tilavuuden. Jos valittu on NONE (EI MITN), tilavuusmuuntoa ei suoriteta. Valitse tarkkailtavan astian tai silin muotoa vastaava astian muoto.
NONE, LINEAR (SUORA), CYLINDER (SYLINTERI), PARABOLIC BOTTOM (PARABOLINEN POHJA), HALF SPHERE BOTTOM (PUOLIYMPYRN MUOTOINEN POHJA), FLAT SLOPED BOTTOM (TASAINEN KALTEVA POHJA), PARABOLIC ENDS (PARABOLISET PT), SPHERE (PALLO), CONICAL BOTTOM (KARTIOMAINEN POHJA), CURVE TABLE (KYRKUVAAJA), LINEAR TABLE (LINEAARINEN KUVAAJA) Oletusarvo: LINEAR

Vaihtoehdot

Jos valitset CURVE TABLE tai LINEAR TABLE, syt tason ja tilavuuden raja-arvot ohjatun toiminnon loppuun suorituksen jlkeen. (Katso parametri 2.6.7.Table 1-8 kyttoppaan laajassa versiossa.) Units (Yksikt)

Mittapn mittausyksikt..

Vaihtoehdot

M, CM, MM, FT, IN Oletusarvo: M

Suomi

Huomautus: Tss esimerkiss kaikki arvot ilmoitetaan oletusarvoisesti


metrein (m).

7ML19985XU81

SITRANS LUT400 (HART)

Sivu FI-21

High Calibration Point (Korkea kalibrointipiste)

Etisyys mittapn viitepisteest korkeaan kalibrointipisteeseen: yleens prosessin ylin taso..

Arvo

Alue: 0,00060,000 Oletusarvo: 0,000

Etisyys mittapn viitepisteest matalaan kalibrointipisteeseen: yleens prosessin perustaso..

Low Calibration Point (Matala kalibrointipiste)

Arvo

Alue: 0,00060,000 Oletusarvo: 60,000

Mritt laitteen reagointinopeuden mittaustulosten muutoksiin valitulla mittausalueella.

Response Rate (Vastenopeus)

Huomautukset:

Vastenopeus voidaan mritt ainoastaan ohjatun pikakytttoiminnon avulla. Aina kun parametreja Fill Rate per Minute (2.3.1.) tai Empty Rate per Minute (2.3.2.) muutetaan ohjatun pikakytttoiminnan pttmisen jlkeen, vastenopeusasetuksen arvot korvautuvat uusilla. Response Rate (Vastenopeus) nytetn aina metrein minuutissa (m/min).
SLOW (0.1 M/MIN) MEDIUM (1.0 M/MIN) FAST (10 M/MIN) Oletusarvo: SLOW (0.1 M/MIN)

Vaihtoehdot

Kyt asetusta, joka on hieman nopeampi kuin tytn tai tyhjentymisen enimmisnopeus (sen mukaan, kumpi on suurempi). Hitaampi asetus on tarkempi, nopeampi asetus sallii suuremmat tasonvaihtelut Dimension A (Mitta A)

Astian pohjan korkeus, kun sen muoto on kartiomainen, pyramidin muotoinen, parabolinen, pallo tai tasainen ja kalteva..

Arvo

Alue: 0,00099,999 Oletusarvo: 0,000

Dimension L (Mitta L)

Vaakasuuntaisen, parabolisen pdyn omaavan astian sylinterimisen osan pituus.

Arvo

Alue: 0,00099,999 Oletusarvo: 0,000

Volume Units (Tilavuusyksikt)

Mritt tilavuuden mittayksikt.


L, USGAL (YHDYSVALTAIN GALLONA), IMPGAL (BRITTILINEN GALLONA),

Vaihtoehdot CUM (KUUTIOMETRI), USER DEFINED (KYTTJN MRITTM)* Suomi


Oletusarvo: L

* Jos valitaan USER DEFINED, arvo on mritettv ohjatun toiminnon loppuun suorituksen jlkeen. Katso parametri 2.6.6.User Defined Unit kyttoppaan laajassa versiossa. Sivu FI-22 SITRANS LUT400 (HART) 7ML19985XU81

Maximum Volume (Enimmistilavuus)

Astian enimmistilavuus. Syt korkeaa kalibrointipistett vastaava astian tilavuus. Jos astian enimmistilavuus on esimerkiksi 8000 L, syt arvoksi 8000.

Arvo

Alue: 0,09999999 Oletusarvo: 100,0

End of QS Volume Wizard (Tilavuusmittauksen ohjatun pikakytttoiminnon loppu)

Jotta ohjattu pikakytttoiminto onnistuu, kaikki muutokset on otettava kyttn.


BACK (TAKAISIN), CANCEL (PERUUTA), FINISH (LOPETA) (Nytt palautuu

Vaihtoehdot kohtaan 1.1 Ohjatun pikakytttoiminnon valikko, kun toiminto on pttynyt


tai peruutetaan. Jos valitaan CANCEL, muutoksia ei tallenneta laitteeseen.)

Siirr pikakynnistysarvot laitteeseen ja palaa ohjelmointivalikkoon painamalla nuoli ALAS (Finish). Palaa sen jlkeen mittaustilaan painamalla kolme kertaa . nuoli VASEMMALLE

1.1.3. QS Flow (Virtausmittauksen ohjattu pikakynnistystoiminto)


Tll ohjatulla pikakytttoiminnolla mritetn yksinkertaisia virtausmittaussovelluksia. [Nkyy ainoastaan LUT430:n (pumppu ja virtaus) ja LUT440:n (OCM, avokanavavirtaus) vastaavasti konfiguroiduissa malleissa.] a) Nuoli OIKEALLE MENU. aktivoi PROGRAM-ohjelmointitilan ja avaa valikkotason 1. MAIN .

b) Siirry valikon kohtaan 1.1.1 painamalla kaksi kertaa nuoli OIKEALLE ja nuoli OIKEALLE .
QUICK START FLOW
TRANSDUCER XRS-5
BACK CANCEL NEXT EDIT

c) Avaa QS Flow (Virtausmittauksen ohjattu pikakytttoiminto) painamalla nuoli ALAS d) Joka kohdassa voit hyvksy oletusarvot ja siirty seuraavaan kohtaan painamalla nuoli ALAS tai siirty muokkaustilaan painamalla nuoli OIKEALLE : valittuna oleva vaihtoehto nkyy korostettuna. e) Vierit halumaasi kohtaan ja tallenna muutos painamalla nuoli OIKEALLE jlkeen painamalla nuoli ALAS f) painamalla nuoli YLS VASEMMALLE . ja jatka sen .

QUICK START FLOW


TRANSDUCER

NO TRANSDUCER XPS-10 XPS-15 XCT-8 XCT-12

XRS-5

Voit milloin tahansa palata taaksepin tai peruuttaa ohjatun toiminnon painamalla nuoli

Start of QS Flow Wizard (Virtausmittauksen ohjatun pikakytttoiminnon kynnistys)

Suomi

Nytt suoritettavan ohjatun toiminnon tyypin.

Vaihtoehdot CANCEL, START

7ML19985XU81

SITRANS LUT400 (HART)

Sivu FI-23

Transducer (Anturi)

Mritt laitteeseen kytketyn Siemens-anturin.


NO TRANSDUCER (EI ANTURIA), XRS-5, XPS-10, XPS-15, XCT-8, XCT-12,

Vaihtoehdot XPS-30, XPS-40, XLT-30, XLT-60, STH


Oletusarvo: NO TRANSDUCER

Temperature Source (Lmptilan lhde)

nen nopeuden stmisess kytettv lmptilalukeman lhde


TRANSDUCER, FIXED TEMPERATURE, EXTERNAL TS-3, AVERAGE OF

Vaihtoehdot SENSORS
Oletusarvo: TRANSDUCER

Fixed Temperature (Kiinte lmptila)

nen nopeuden stmisess kytettv lmptilalukeman lhde

Arvo

Alue: -100,0+150,0 C Oletusarvo: +20,0 C

Tm parametri nkyy vain, jos FIXED TEMPERATURE on valittu kohdassa Temperature Source (Lmptilan lhde). Primary Measuring Device (Ensisijainen mittauslaite)

Mritt sovelluksessa kytettvn ensisijaisen mittauslaitteen (PMD).

Vaihtoehdot

EXPONENTIAL DEVICES (EKSPONENTIAALISET LAITTEET), RECTANGULAR FLUME BS-3680 (SUORAKULMAINEN MITTAKOURU BS-3680), ROUND NOSE HORIZONTAL CR. BS-3680 (PYREKRKINEN VAAKASUUNTAINEN KYNNYSPATO BS-3680), TRAPEZOIDAL FLUME BS-3680 (CIPOLETTIN MITTAKOURU BS-3680), U-FLUME BS-3680 (U-MITTAKOURU BS-3680), FINITE CREST WEIR BS-3680 (RAJALLINEN KYNNYSPATO BS-3680), THIN PLATE RECT. WEIR BS-3680 (TERHARJAINEN SUORAKULMAINEN MITTAPATO BS-3680), THIN PLATE V-NOTCH WEIR BS-3680 (TERHARJAINEN KOLMION MUOTOINEN MITTAPATO BS-3680), RECT. WEIR CONTRACTED (SIVUILTA KURISTUVA SUORAKULMAINEN MITTAPATO), ROUND PIPE (PYRE PUTKI), PALMER BOWLUS FLUME (PALMER BOWLUS -MITTAKOURU), H-FLUME (HMITTAKOURU), OTHER (MUU)* Oletusarvo: EXPONENTIAL DEVICES

* Vaihtoehdoksi mrittyy OTHER, jos ohjattu toimintoa on suoritettu aiemmin HARTohjelmistotykalun avulla (esim. SIMATIC PDM) ja laitteen tilaksi on mritetty OFF (POIS) tai UNIVERSAL HEAD VS. FLOW (YLEISKORKEUS V. VIRTAUS). Jos kyseess on ensimminen mrityskerta, PMD voidaan asettaa ainoastaan tilaan ilman laskentaa (OFF (POIS)) tai linearisointia varten (UNIVERSAL HEAD VS. FLOW) HARTohjelmistotykaluilla (SIMATIC PDM, AMS, FC375/475). Katso listiedot kyttoppaan laajasta versiosta. Method of Flow Calculation (Virtauksen laskentamenetelm)

Mritt virtauksen laskentamenetelmn.

Vaihtoehdot
Units (Yksikt)

ABSOLUTE (ABSOLUUTTINEN), RATIOMETRIC (SUHDEMITTAUS) Oletusarvo: ABSOLUTE

Mittapn mittausyksikt..

Suomi

Vaihtoehdot

M, CM, MM, FT, IN Oletusarvo: M

Huomautus: Tss esimerkiss kaikki arvot ilmoitetaan oletusarvoisesti


metrein (m).

Sivu FI-24

SITRANS LUT400 (HART)

7ML19985XU81

High Calibration Point (Korkea kalibrointipiste)

Etisyys mittapn viitepisteest korkeaan kalibrointipisteeseen: yleens prosessin ylin taso..

Arvo

Alue: 0,00060,000 Oletusarvo: 0,000

Low Calibration Point (Matala kalibrointipiste)

Etisyys mittapn viitepisteest matalaan kalibrointipisteeseen: yleens prosessin perustaso..

Arvo

Alue: 0,00060,000 Oletusarvo: 60,000

Mritt laitteen reagointinopeuden mittaustulosten muutoksiin valitulla mittausalueella.

Response Rate (Vastenopeus)

Huomautukset:

Vastenopeus voidaan mritt ainoastaan ohjatun pikakytttoiminnon avulla. Aina kun parametreja Fill Rate per Minute (2.3.1.) tai Empty Rate per Minute (2.3.2.) muutetaan ohjatun pikakytttoiminnan pttmisen jlkeen, vastenopeusasetuksen arvot korvautuvat uusilla. Response Rate (Vastenopeus) nytetn aina metrein minuutissa (m/min).
SLOW (0.1 M/MIN) MEDIUM (1.0 M/MIN) FAST (10 M/MIN) Oletusarvo: SLOW (0.1 M/MIN)

Vaihtoehdot

Kyt asetusta, joka on hieman nopeampi kuin tytn tai tyhjentymisen enimmisnopeus (sen mukaan, kumpi on suurempi). Hitaampi asetus on tarkempi, nopeampi asetus sallii suuremmat tasonvaihtelut. Calculation factors (Laskentakertoimet):

Huomautukset:

Seuraavat viisi parametria nkyvt ohjatussa toiminnossa kulloinkin aina edell valitun ensisijaisen mittauslaitteen mukaan. Nit parametreja kytetn virtauksen laskentakaavassa (katso yksityiskohtaiset tiedot kyttoppaan laajasta versiosta). Flow Exponent (Virtauksen eksponentti) (PMD = EXPONENTIAL DEVICES) (ENSISIJAINEN MITTAUSLAITE = EKSPONTENTIAALISET LAITTEET)

Virtauksen laskentakaava eksponentti.

Arvo

Alue: -999,000+9999,000 Oletusarvo: 1,550

Suomi

7ML19985XU81

SITRANS LUT400 (HART)

Sivu FI-25

K Factor (K-kerroin) (PMD = EXPONENTIAL DEVICES) (ENSISIJAINEN MITTAUSLAITE = EKSPONENTIAALISET LAITTEET)

Yksinomaan eksponentiaalisen laitteen absoluuttisessa laskennassa virtauksen laskentakaavassa kytettv vakio.

Arvo

Alue: -999,000+9999,000 Oletusarvo: 1,000

V-Notch Angle (Kolmion kulma) (PMD = THIN PLATE V-NOTCH WEIR) (ENSISIJAINEN MITTAUSLAITE = TERHARJAINEN KOLMION MUOTOINEN MITTAPATO)

Kolmion kulmaa kytetn virtauksen laskentakaavassa.

Arvo

Alue: 25,00095,000 Oletusarvo: 25,000

Slope (Kaltevuus) (PMD = TRAPEZOIDAL FLUME or ROUND PIPE) (ENSISIJAINEN MITTAUSLAITE = CIPOLETTIN MITTAKOURU tai PYRE PUTKI)

Virtauksen laskentakaavassa kytettv virtauksen kaltevuus.

Arvo

Alue: -999,000+9999,000 Oletusarvo: 0,000

Roughness Coefficient (Karheuskerroin) (PMD = ROUND PIPE) (ENSISIJAINEN MITTAUSLAITE = PYRE PUTKI)

Virtauksen laskentakaavassa kytettv karheuskerroin.

Arvo

Alue: -999,000+9999,000 Oletusarvo: 0,000

Suomi
Sivu FI-26

SITRANS LUT400 (HART)

7ML19985XU81

PMD Dimensions (Ensisijaisen mittauslaitteen mitat)

Huomautukset:

Jokaiselle ensisijaiselle mittauslaitteelle on Exponential Devices (Eksponentiaaliset laitteet) ja Other (Muu) -vaihtoehtoja lukuun ottamatta sytettv nelj mittaa. Ohjattu toiminta pyyt kutakin valitussa ensisijaisessa mittauslaitteessa tarvittavaa mittaa, ja nyttn tulee vastaava mittalaitteen mitan nimi. Ohjatun toiminnon mitan nimi (parametrivalikon tiedot)
APPROACH WIDTH B (TULOLEVEYS B) (2.15.4.5. OCM Dimension 1) THROAT WIDTH B (NIELUN LEVEYS B) (2.15.4.6. OCM Dimension 2) HUMP HEIGHT P (PORTAAN KORKEUS P) (2.15.4.7. OCM Dimension 3) THROAT LENGTH L (NIELUN PITUUS L) (2.15.4.8. OCM Dimension 4)

Valittu PMD

Rectangular Flume BS-3680 (Suorakulmainen mittakouru BS-3680)

Round Nose Horizontal Crest Weir BS-3680 (Pyrekrkinen vaakasuuntainen kynnyspato BS-3680)
CREST WIDTH B (PADONHARJAN LEVEYS B) (2.15.4.5. OCM Dimension 1) CREST HEIGHT P (PADONHARJAN KORKEUS P) (2.15.4.6. OCM Dimension 2) CREST LENGTH L (PADONHARJAN PITUUS L) (2.15.4.7. OCM Dimension 3)

Trapezoidal Flume BS-3680 (Cipolettin mittauskouru BS-3680)


APPROACH WIDTH B (2.15.4.5. OCM Dimension 1) (TULOLEVEYS B) THROAT WIDTH B (2.15.4.6. OCM Dimension 2) (NIELUN LEVEYS B) HUMP HEIGHT P (2.15.4.7. OCM Dimension 3) (PORTAAN KORKEUS P) THROAT LENGTH L (2.15.4.8. OCM Dimension 4) (NIELUN PITUUS L)

U-Flume BS-3680 (U-mittakouru BS-3680)


APPROACH DIAMETER DA (TULOHALKAISIJA DA) (2.15.4.5. OCM Dimension 1) THROAT DIAMETER D (NIELUN HALKAISIJA D) (2.15.4.6. OCM Dimension 2) HUMP HEIGHT P (PORTAAN KORKEUS P) (2.15.4.7. OCM Dimension 3) THROAT LENGTH L (NIELUN PITUUS L) (2.15.4.8. OCM Dimension 4)

Finite Crest Weir BS-3680 (Rajallinen kynnyspato BS-3680)


CREST WIDTH B (PADONHARJAN LEVEYS B) (2.15.4.5. OCM Dimension 1) CREST HEIGHT P (PADONHARJAN KORKEUS P) (2.15.4.6. OCM Dimension 2) CREST LENGTH L (PADONHARJAN PITUUS L) (2.15.4.7. OCM Dimension 3)

Thin Plate Rectangular Weir BS-3680 (Terharjainen suorakulmainen mittapato BS-3680)


APPROACH WIDTH B (TULOLEVEYS B) (2.15.4.5. OCM Dimension 1) CREST WIDTH B (PADONHARJAN LEVEYS B) (2.15.4.6. OCM Dimension 2) CREST HEIGHT P (PADONHARJAN KORKEUS P) (2.15.4.7. OCM Dimension 3)

Rectangular Weir Contracted (Sivuilta kuristuva suorakulmainen mittapato)


CREST WIDTH B (PADONHARJAN LEVEYS B) (2.15.4.5. OCM Dimension 1) (jatkuu seuraavalla sivulla)

Suomi

7ML19985XU81

SITRANS LUT400 (HART)

Sivu FI-27

Valittu PMD (jatkoa)

Ohjatun toiminnon mitan nimi (parametrivalikon tiedot)


PIPE INSIDE DIAMETER D (PUTKEN SISHALKAISIJA D) (2.15.4.5. OCM Dimension 1)

Round Pipe (Pyre putki)

Palmer Bowlus Flume (Palmer Bowlus -mittakouru)


MAXIMUM FLUME WIDTH HMAX (MITTAKOURUN ENIMMISLEVEYS HMAX) (2.15.4.5. OCM Dimension 1)

H-Flume (H-mittakouru)
MAXIMUM LISTED HEAD HMAX (TALLENNETTU ENIMMISKORKEUS HMAX) (2.15.4.5. OCM Dimension 1)

Maximum Head (Enimmiskorkeus)

Mittauslaitteeseen liittyv suurin tasoarvo.

Arvo

Alue: 0,00060,000 Oletusarvo: 60,000

Zero Head Offset (0-korkeuden poikkeama)

Erotus (positiivinen tai negatiivinen) Low Calibration Point (Matala kalibrointipiste) eli Matalan kalibrointipisteen ja 0-korkeuden vlill (taso 0-virtauksella).

Arvo

Alue: -60,000+60,000 Oletusarvo: 0,000

Flowrate Units (Virtausnopeuden yksikt)

Virtaaman nytss kytettv tilavuusyksikt.


L/S, L/MIN, CUFT/S (KUUTIOJALKAA/S), CUFT/D (KUUTIOJALKAA/PIV), GAL/MIN, GAL/D (US GALLONAA/PIV, IMPGAL/MIN (ISONBRITANNIAN GALLONAA/MIN), IMPGAL/D ((ISON-BRITANNIAN GALLONAA/PIV)), CUM/H (KUUTIOMETRI/H), CUM/D (KUUTIOMETRI/PIV), USER DEFINED (KYTTJN MRITTM) * Oletusarvo: L/S

Vaihtoehdot

* Jos valitaan USER DEFINED, arvo on mritettv ohjatun toiminnon loppuun suorituksen jlkeen. Katso parametri 2.6.6.User Defined Unit kyttoppaan laajassa versiossa. Maximum Flow at 20 mA (Enimmisvirtaus 20 mA:n virralla)

Enimmisvirtaama.

Arvo

Alue: -9999999999 Oletusarvo: 100

Flowrate Decimal (Virtaaman desimaalit)

Nytettvien desimaalien enimmismr.


NO DIGITS (EI DESIMAALEJA), 1 DIGIT, 2 DIGITS, 3 DIGITS (1, 2, 3

Vaihtoehdot NUMEROA)
Oletusarvo: NO DIGITS

Suomi

Low Flow Cutoff (Hitaan virtauksen raja-arvo)

Keskeytt totalisaattorin toiminnan, kun energiakorkeuden taso on raja-arvon korkeudella tai alempi.

Arvo
Sivu FI-28

Alue: 0,00060,000 Oletusarvo: 0,000

SITRANS LUT400 (HART)

7ML19985XU81

End of QS Flow Wizard (Virtausmittauksen ohjatun pikakytttoiminnon loppu)

Jotta ohjattu pikakytttoiminto onnistuu, kaikki muutokset on otettava kyttn.


BACK (TAKAISIN), CANCEL (PERUUTA), FINISH (LOPETA) (Nytt palautuu Vaihtoehdot kohtaan 1.1.1 Ohjatun pikakytttoiminnon valikko, kun toiminto on pttynyt tai peruutetaan. Jos valitaan CANCEL, muutoksia ei tallenneta laitteeseen.)

Siirr pikakynnistysarvot laitteeseen ja palaa ohjelmointivalikkoon painamalla nuoli ALAS (Finish). Palaa sen jlkeen mittaustilaan painamalla kolme kertaa . nuoli VASEMMALLE

Huomautus: Parhaan mahdollisen tarkkuuden varmistamiseksi suosittelemme 2.15.2.Auto Zero Head -toimintoa ohjatun toiminnon valmiiksi suorittamisen jlkeen.
Katso yksityiskohtaiset ohjeet kyttoppaan laajasta versiosta.

1.2. Pumpun ohjaus


Tll ohjatulla toiminnolla mritetn pumput, mikli niit kytetn sovelluksessasi. Varmista, ett suoritat ensin sopivan ohjatun pikakynnistystoiminnon. a) Nuoli OIKEALLE MENU. aktivoi PROGRAM-ohjelmointitilan ja avaa valikkotason 1. MAIN . ja nuoli

b) Siirry valikon kohtaan 1.1 painamalla nuoli OIKEALLE

c) Avaa Pump Control (Pumpun ohjaus) painamalla nuoli ALAS OIKEALLE .

d) Joka kohdassa voit hyvksy oletusarvot ja siirty seuraavaan kohtaan painamalla tai siirty muokkaustilaan painamalla nuoli OIKEALLE nuoli ALAS oleva vaihtoehto nkyy korostettuna. e) Vierit halumaasi kohtaan ja tallenna muutos painamalla nuoli OIKEALLE sen jlkeen painamalla nuoli ALAS f) . tai peruuttaa : valittuna ja jatka

Voit milloin tahansa palata taaksepin painamalla nuoli YLS ohjatun toiminnon painamalla nuoli VASEMMALLE .

Start of Wizard - Pump Control (Pumpun ohjauksen ohjatun toiminnon kynnistys)

Nytt suoritettavan ohjatun toiminnon tyypin.

Vaihtoehdot

CANCEL, START

Number of Pumps (Pumppujen mr)

Valitse pumpun ohjauksen avulla kytettvien pumppujen lukumr.

Vaihtoehdot

NONE (EI YHTN), 2 Oletusarvo: NONE

Suomi

Jos asetukseksi valitaan NONE, pumpun ohjaus kytkeytyy pois kytst.

7ML19985XU81

SITRANS LUT400 (HART)

Sivu FI-29

Relay Pump 1 (Pumpun 1 rele)

Valitsee Pumpulle 1 osoitetun releen.

Vaihtoehdot

RELAY 2 (RELE 2), RELAY 3 (RELE 3) Oletusarvo: RELAY 2

Relay Pump 2 (Pumpun 2 rele)

Pelkstn katsottavissa. Valitsee automaattisesti Pumpulle 2 osoitetun releen edellisess vaiheessa Pumpulle 1 valitun releen perusteella.
Jos Pumpun 1 rele = RELAY 2, niin Pumpun 2 rele = RELAY 3 Vaihtoehdot (pelkstn Jos Pumpun 1 rele = RELAY 3, niin Pumpun 2 rele = RELAY 2 katsottavissa)

Pump Control Mode (Pumpun ohjaustapa)

Mritt releen laukaisuun kytettvn ohjausalgoritmin.

LUT420 -tasonohjain:
ALTERNATE DUTY ASSIST, ALTERNATE DUTY BACKUP

LUT430 pumppu- ja virtausohjain:


ALTERNATE DUTY ASSIST, ALTERNATE DUTY BACKUP , SERVICE RATIO DUTY ASSIST, SERVICE RATIO DUTY BACKUP , FIXED DUTY Mallikohtaisesti ASSIST, FIXED DUTY BACKUP tuetut vaihtoehdot LUT440 erittin tarkka avokanavirtauksen ohjaus: ALTERNATE DUTY ASSIST, ALTERNATE DUTY BACKUP , SERVICE RATIO DUTY ASSIST, SERVICE RATIO DUTY BACKUP , FIXED DUTY ASSIST, FIXED DUTY BACKUP Oletus (kaikki mallit): ALTERNATE DUTY ASSIST

Service Ratio Pump 1 (Pumpun 1 kyttsuhde)

Valitsee pumpun kytn kyntiaikasuhteen eik viimeksi kytetyn pumpun perusteella.

Arvo

Alue: 0255 Oletusarvo: 1

Tm parametri nkyy vain, jos Pump Control Mode (Pumpun ohjaustapa) asetukseksi on valittu Service Ratio (Kyttsuhde) -algoritmi. Service Ratio Pump 2 (Pumpun 2 kyttsuhde)

Valitsee pumpun kytn kyntiaikasuhteen eik viimeksi kytetyn pumpun perusteella..

Arvo

Alue: 0255 Oletusarvo: 1

Tm parametri nkyy vain, jos Pump Control Mode -asetukseksi on valittu Service Ratio -algoritmi.

Suomi
Sivu FI-30

SITRANS LUT400 (HART)

7ML19985XU81

Run Time Relay 2 (Releen 2 kyntiaika)

Mritt releen 2 pumpun kyntiajan tunteina.

Arvo

Alue: 0999999 Oletusarvo: 0

Kyt uusien pumppujen yhteydess oletusarvoa tai mrit kytetyille pumpuille niiden kokonaiskyntimr tll arvolla. (Tm parametri nkyy vain, jos Pump Control Mode -asetukseksi on valittu Service Ratio -algoritmi.) Run Time Relay 3 (Releen 3 kyntiaika)

Mritt releen 3 pumpun kyntiajan tunteina.

Arvo

Alue: 0999999 Oletusarvo: 0

Kyt uusien pumppujen yhteydess oletusarvoa tai mrit kytetyille pumpuille niiden kokonaiskyntimr tll arvolla. (Tm parametri nkyy vain, jos Pump Control Mode -asetukseksi on valittu Service Ratio -algoritmi.) ON Setpoint Pump 1 (Pumpun 1 kynnistyksen laukaisupiste)

Pumpun 1 kynnistymistaso, joka on mritetty kohdassa 2.1.1.Units.

Arvo

Alue: 0,00099999,000 Oletusarvo: 0,000

ON Setpoint Pump 2 (Pumpun 2 kynnistyksen laukaisupiste)

Pumpun 2 kynnistymistaso, joka on mritetty kohdassa 2.1.1.Units

Arvo

Alue: 0,00099999,000 Oletusarvo: 0,000

OFF Setpoint Pump 1 (Pumpun 1 pysytyksen laukaisupiste)

Pumpun 1 pyshtymistaso, joka on mritetty kohdassa 2.1.1.Units.

Arvo

Alue: 0,00099999,000 Oletusarvo: 0,000

OFF Setpoint Pump 2 (Pumpun 2 pysytyksen laukaisupiste)

Pumpun 2 pyshtymistaso, joka on mritetty kohdassa 2.1.1.Units.

Arvo

Alue: 0,00099999,000 Oletusarvo: 0,000

End of Wizard - Pump Control (Pumpun ohjauksen ohjatun toiminnon loppu)

Jotta ohjattu toiminto onnistuu, kaikki muutokset on otettava kyttn.


BACK, CANCEL, FINISH (Nytt palautuu Pumpun ohjausvalikkoon, kun

Vaihtoehdot toiminto on pttynyt tai peruutetaan. Jos valitaan CANCEL, muutoksia ei


tallenneta laitteeseen.)

Siirr laitteeseen ja palaa ohjelmointivalikkoon painamalla nuoli ALAS

(Finish). .

Suomi

Palaa sen jlkeen mittaustilaan painamalla kolme kertaa nuoli VASEMMALLE

7ML19985XU81

SITRANS LUT400 (HART)

Sivu FI-31

SITRANS LUT400 -tiedonsiirto: HART


SITRANS LUT400 voidaan konfiguroida HART-verkon yli kyttmll Emersonin valmistamaa HART Communicator 375/475 -taskutieturia (katso yksityiskohtaiset ohjeet kyttoppaan laajasta versiosta) tai ohjelmistopakettia. Suosittelemme Siemensin valmistamaa SIMATIC Process Device Manager (PDM) -ohjelmistopakettia. Yksityiskohtaiset ohjeet on esitetty sovellusoppaassa Working with HART (Tyskentely HARTverkossa), joka voidaan ladata tuotesivulta sivustostamme. Siirry sivulle www.siemens.com/ sitransLUT400 ja kohtaan Support (Tuki) ja napsauta Application Guides (Sovellusoppaat).

Etiskytt
SITRANS LUT400 tukee useita ohjelmistotykaluja, joiden avulla laitetta voidaan kytt etistiedonsiirron avulla: SIMATIC PDM -ohjelmisto pc-tietokoneessa Emerson AMS Device Manager -laitehallintaohjelmisto pc-tietokoneessa Verkkoselain pc-tietokoneessa Field Device Tool (FDT) -kenttvyltykalu pc-tietokoneessa Field Communicator 375/475 (FC375/FC475). Seuraavassa on esitetty perustiedot niden tykalujen kyttmisest SITRANS LUT400:n yhteydess. Listietoja eri tykaluista on saatavissa SITRANS LUT400 (HART) -tiedonsiirtooppaasta.

SIMATIC PDM
SIMATIC PDM on ohjelmistopaketti, jolla otetaan kyttn ja kunnossapidetn SITRANS LUT400:aa ja muita prosessilaitteita. Voit lukea listietoja osoitteesta www.siemens.com/ simatic-pdm. Tarkista verkkosivustomme tukisivulta, onko Sinulla varmasti SIMATIC PDM:n uusin versio ja uusimmat Service Pack (SP) - ja hot fix (HF) -tiedostot. Siirry sivulle http://support.automation.siemens.com/WW/ llisapi.dll?func=cslib.csinfo&lang=en&objiD=10806857&subtype=133100

Pivit shkinen laitemritys (EDD)


Laitemritys lytyy laiteluettelosta (Device Catalog) kohdasta Sensors/Level/Echo/Siemens AG/SITRANS LUT400. (Shkinen laitemritys on suunniteltu yhteensopivaksi tulevien versioiden kanssa.) Oikean shkisen laitemrityksen lytmisess on avuksi, kun muistaa, ett laitemrityksen version ja laitteessa olevan laiteohjelmistoversion ensimmisten ja toisten (pisteen jlkeisten) numeroiden tulee vastata toisiaan (esimerkiss lihavoidut ensimmiset ja toiset numerot: 1.00.00-02).
Laitemrityksen versionumeron on oltava sama kuin laitteessa olevan laiteohjelmiston versionumero. Tmn voi tarkistaa kohdasta SITRANS LUT400 HART > Identification (Tunnistus)> Device (Laite).
toisiaan vastaava laiteohjelmistoversio ja EDD-versio

SIMATIC PDM:n uuden version asentaminen edellytt uusimman Service Pack (SP) - ja uusimman hot fix (HF) -tiedoston kytt.

Uuden laitemrityksen asentaminen Suomi

Siirry sivulle www.siemens.com/sitransLUT400 > Support (Tuki) > Software Downloads (Ohjelmistolataukset) ja lataa uusin shkinen laitemritys tuotesivustolta sivustostamme. Tallenna tiedostot tietokoneellesi ja pura pakattu tiedosto helppopsyiseen kansioon. Kynnist SIMATIC PDM Manager Device Catalog (Hallintalaiteluettelo). Siirry puretun EDD-tiedoston sisltvn kansioon ja valitse kyseinen kansio. SITRANS LUT400 (HART) 7ML19985XU81

Sivu FI-32

Uuden laitteen kokoonpanoasetukset


1) 2) Tarkista, ett sinulla on uusin laitemritys, ja pivit se tarvittaessa (katso edellUuden laitemrityksen asentaminen). Konfiguroi SITRANS LUT400 omaa sovellustasi vastaavalla ohjatulla pikakynnistystoiminnolla.

Ohjatut pikakytttoiminnot SIMATIC PDM:n avulla


Huomautukset:
Ohjatun pikakytttoiminnon asetukset ovat sidoksissa toisiinsa, ja muutokset tulevat voimaan vasta, kun napsautat kohtaa Apply and Transfer (Ota kyttn ja siirr) viimeisess vaiheessa. Napsauta BACK (TAKAISIN), jos haluat palata taaksepin tarkistamaan jonkin asetuksen, tai CANCEL (PERUUTA), jos haluat poistua Quick Start -pikakynnistystoiminnosta. Avaa SIMATIC PDM:n Device (Laite) -valikko ja valitse ohjattu pikakynnistystoiminto omaa sovellustasi varten (Level, Volume, Volume-Linearization (Tilavuuden linearisointi), Flow).

Pikakytttoiminnot

Kytt AMS-laitehallintaohjelmiston avulla


AMS Device Manager on ohjelmistopaketti, jolla otetaan kyttn ja kunnossapidetn SITRANS LUT400:AA ja muita prosessilaitteita. Voit lukea listietoja osoitteesta http://www.emersonprocess.com/AMS/.

Shkinen laitemritys (EDD)


SITRANS LUT400 edellytt AMS Device Manager -ohjelmistoversiolle 10.5 suunnitellun EDDlaitemrityksen.

Uuden laitteen konfigurointi AMS Device Managerin avulla


1) Laitemritys lytyy Device Catalog -laiteluettelosta kohdasta Sensors/Level/Echo/Siemens AG/SITRANS LUT400. Tarkista tuotesivulta verkkosivustostamme osoitteessa www.siemens.com/sitransLUT400 kohdasta Support->Software Downloads (TukiOhjelmistolataukset), ett sinulla on varmasti AMS Device Managerin laitemrityksen (EDD) uusin versio. Tallenna tiedostot tietokoneellesi ja pura pakattu tiedosto helppopsyiseen kansioon. Kynnist AMS Device Manager Add Device Type (Lis laitetyyppi), selaa puretun EDD-tiedoston kohdalle ja valitse se. (Sovellusopas HART-laitteiden asetusten mrittmiseksi AMS Device Managerilla on ladattavissa sivustomme tuotesivulta kohdasta Support (Tuki).) Napsauta kohdasta Device Connection View -laiteliitntnytss hiiren kakkospainikkeella laitteen kuvaketta ja valitse Scan Device (Laitteen skannaus), jolloin parametrit ladataan laitteesta. Avaa aloitusnytt kaksoisnapsauttamalla laitteen kuvaketta. Aloitusnytss nkyvt laitteen tunnistetiedot. Nytn vasemmalla puolella on navigointi-ikkuna. Konfiguroi SITRANS LUT400 asianmukaisella ohjatulla pikakynnistystoiminnolla. SITRANS LUT400 (HART) Sivu FI-33

2) 3)

4) 5) 6)

Suomi

7ML19985XU81

Kytt FC375/475 (HART) -yhteyslaitteella


FC375/FC475 HART Communicator on helppokyttinen taskutieturi (tiedonsiirtolaite). Se sislt yleisen tuen muille HART-laitteille, esim. SITRANS LUT400. FC375/475:n valikkorakenne muistuttaa hyvin paljon AMS Device Manager -ohjelmistoa.

Shkinen laitemritys (EDD)


SITRANS LUT400 edellytt FC375/FC475:lle suunniteltua elektronista laitemrityst (EDD).

Uuden laitteen konfigurointi FC375/FC475:n avulla


1) Tarkista tuotesivulta verkkosivustostamme osoitteessa www.siemens.com/sitransLUT400 kohdasta Support->Software Downloads (Tuki-Ohjelmistolataukset), ett sinulla on varmasti FC375/FC475:n EDD-laitemrityksen uusin versio. Tallenna tiedostot tietokoneellesi ja pura pakattu tiedosto helppopsyiseen kansioon. Asenna EDD suorittamalla Emersonin ohjelmisto Field Communicator Easy Upgrade Utility (FC-tiedonsiirtolaitteen pikapivitysohjelmisto). Konfiguroi SITRANS LUT400 omaa sovellustasi vastaavalla ohjatulla pikakynnistystoiminnolla.

2) 3) 4)

Kytt FDT-kenttvyltykalun (Field Device Tool) avulla


FDT on useissa eri ohjelmistopaketeissa kytetty vakiomuotoinen tykalu, jolla otetaan kyttn ja yllpidetn kenttlaitteita, kuten SITRANS LUT400. Markkinoilla on saatavissa kaksi kenttvyltykalua: PACTware ja Fieldcare. FDT muistuttaa hyvin paljon PDM-ohjelmistoa. Jotta kenttlaite voidaan konfiguroida kenttvyltykalun avulla, tarvitaan laitteen DTM-ajuriohjelma (Device Type Manager) . Siemens-mittauslaitteissa kytetn SITRANS DTM -ohjelmistoa ja SITRANS DTM ohjelmistoa varten kirjoitettua mittauslaitteen EDD-laitemrityst. 1) Asenna ensin SITRANS DTM jrjestelmsi. Sen voi ladata osoitteesta http://www.siemens.com/sitransdtm. Napsauta kohtaa Support (Tuki) ja sen jlkeen Software downloads (Ohjelmistolataukset). Asenna SITRANS LUT400 (HART) -laitemritys (EDD), joka on tarkoitettu SITRANS DTM -ohjelmistolle. Se voidaan ladata tuotesivulta sivustostamme: Siirry sivulle www.siemens.com/sitransLUT400 kohtaan Support (Tuki) ja napsauta Software Downloads (Ohjelmistolataukset).

2)

Uuden laitteen konfigurointi FDT:n avulla


Sovellusopas voidaan ladata tuotesivulta sivustostamme kohdasta Support (Tuki).

Huolto ja kunnossapito
SITRANS LUT400 ei normaaleissa kyttolosuhteissa vaadi huoltoa tai puhdistamista.

Laitteen korjaaminen ja vastuuvapauslauseke


Katso listietoja takakannen sissivulta.

Suomi
Sivu FI-34

SITRANS LUT400 (HART)

7ML19985XU81

Kytkentmritykset asennettaessa rjhdysvaarallisiin tiloihin


Rjhdysvaarallisten tilojen asennuksia varten on olemassa seuraavat johdotusvaihtoehdot: Plyss sytyttmtn johdotus (kansainvlinen/Eurooppa) Sytyttmtn johdotusmenetelm (Kanada) Tarkista ja varmista kaikissa tapauksissa laitekilvest laitteen hyvksyntluokitus.

1. Plyss sytyttmtn johdotus (kansainvlinen/Eurooppa)


SITRANS LUT420 7MLxxxx-xxxxx-xxxx Serial No.: GYZ / B1034567 0518 KCC-REM-S49 Power Rating: 100 230V 15% SITRANSLUT400 50/60 Hz, 36 VA (10 W) Contact Rating: 1A/5A @ 250V , Non-Inductive Operating Temperature: 20C to 50C Enclosure: IP65 / TYPE 4X / NEMA 4X II 3D Ex tc IIIC T70C Dc Ex tc IIIC T70C Dc SIRA 12ATEX9143X IECEx SIR 12.0057X Warning: Possible static hazard. Do not rub or clean on site with a dry cloth. Warning: Do not open when energized or when an explosive dust atmosphere is present. Siemens Milltronics Process Instruments, Peterborough Assembled in Canada with domestic and imported parts SITRANS LUT430 7MLxxxx-xxxxx-xxxx Serial No.: GYZ / B1034567 0518 KCC-REM-S49 Power Rating: 100 230V 15% SITRANSLUT400 50/60 Hz, 36 VA (10 W) Contact Rating: 1A/5A @ 250V , Non-Inductive Operating Temperature: 20C to 50C Enclosure: IP65 / TYPE 4X / NEMA 4X II 3D Ex tc IIIC T70C Dc Ex tc IIIC T70C Dc SIRA 12ATEX9143X IECEx SIR 12.0057X Warning: Possible static hazard. Do not rub or clean on site with a dry cloth. Warning: Do not open when energized or when an explosive dust atmosphere is present. Siemens Milltronics Process Instruments, Peterborough Assembled in Canada with domestic and imported parts SITRANS LUT440 7MLxxxx-xxxxx-xxxx Serial No.: GYZ / B1034567 0518 KCC-REM-S49 Power Rating: 100 230V 15% SITRANSLUT400 50/60 Hz, 36 VA (10 W) Contact Rating: 1A/5A @ 250V , Non-Inductive Operating Temperature: 20C to 50C Enclosure: IP65 / TYPE 4X / NEMA 4X II 3D Ex tc IIIC T70C Dc Ex tc IIIC T70C Dc SIRA 12ATEX9143X IECEx SIR 12.0057X Warning: Possible static hazard. Do not rub or clean on site with a dry cloth. Warning: Do not open when energized or when an explosive dust atmosphere is present. Siemens Milltronics Process Instruments, Peterborough Assembled in Canada with domestic and imported parts

SITRANS LUT420 7MLxxxx-xxxxx-xxxx Serial No.: GYZ / B1034567 0518 KCC-REM-S49 Power Rating: 10 32V , 10W SITRANSLUT400 Contact Rating: 1A/5A @ 250V , Non-Inductive Operating Temperature: 20C to 50C Enclosure: IP65 / TYPE 4X / NEMA 4X

SITRANS LUT430 7MLxxxx-xxxxx-xxxx Serial No.: GYZ / B1034567 0518 KCC-REM-S49 Power Rating: 10 32V , 10W SITRANSLUT400 Contact Rating: 1A/5A @ 250V , Non-Inductive Operating Temperature: 20C to 50C Enclosure: IP65 / TYPE 4X / NEMA 4X

SITRANS LUT440 7MLxxxx-xxxxx-xxxx Serial No.: GYZ / B1034567 0518 KCC-REM-S49 Power Rating: 10 32V , 10W SITRANSLUT400 Contact Rating: 1A/5A @ 250V , Non-Inductive Operating Temperature: 20C to 50C Enclosure: IP65 / TYPE 4X / NEMA 4X

II 3D Ex tc IIIC T70C Dc Ex tc IIIC T70C Dc SIRA 12ATEX9143X IECEx SIR 12.0057X Warning: Possible static hazard. Do not rub or clean on site with a dry cloth. Warning: Do not open when energized or when an explosive dust atmosphere is present.

II 3D Ex tc IIIC T70C Dc Ex tc IIIC T70C Dc SIRA 12ATEX9143X IECEx SIR 12.0057X Warning: Possible static hazard. Do not rub or clean on site with a dry cloth. Warning: Do not open when energized or when an explosive dust atmosphere is present.

II 3D Ex tc IIIC T70C Dc Ex tc IIIC T70C Dc SIRA 12ATEX9143X IECEx SIR 12.0057X Warning: Possible static hazard. Do not rub or clean on site with a dry cloth. Warning: Do not open when energized or when an explosive dust atmosphere is present.

Siemens Milltronics Process Instruments, Peterborough Assembled in Canada with domestic and imported parts

Siemens Milltronics Process Instruments, Peterborough Assembled in Canada with domestic and imported parts

Siemens Milltronics Process Instruments, Peterborough Assembled in Canada with domestic and imported parts

Laitekilvess luetellut ATEX-sertifikaatit voidaan ladata tuotesivulta sivustostamme osoitteesta www.siemens.com/sitransLUT400. Laitekilvess mainittu IECEx-sertifikaatti on tarkistettavissa IECEx:n verkkosivustossa. Mene sivulle http://iecex.iec.ch ja napsauta kohtaa Ex Equipment Certificates of Conformity . Syt sen jlkeen sertifikaatin numero IECEx SIR 12.0057X. Noudata johdotuksessa aina paikallisia mryksi. Katso mys edell mainittu ATEX-sertifikaatti.

Suomi

7ML19985XU81

SITRANS LUT400 (HART)

Sivu FI-35

2. Sytyttmtn johdotusmenetelm (Kanada)


SITRANS LUT420 7MLxxxx-xxxxx-xxxx Serial No.: GYZ / B1034567 Power Rating: 100 230V 15% 50/60 Hz, 36 VA (10 W) Contact Rating: 1A/5A @ 250V , Non-Inductive Operating Temperature: 20C to 50C Enclosure: IP65 / TYPE 4X / NEMA 4X SITRANS LUT430 7MLxxxx-xxxxx-xxxx Serial No.: GYZ / B1034567 Power Rating: 100 230V 15% 50/60 Hz, 36 VA (10 W) Contact Rating: 1A/5A @ 250V , Non-Inductive Operating Temperature: 20C to 50C Enclosure: IP65 / TYPE 4X / NEMA 4X SITRANS LUT440 7MLxxxx-xxxxx-xxxx Serial No.: GYZ / B1034567 Power Rating: 100 230V 15% 50/60 Hz, 36 VA (10 W) Contact Rating: 1A/5A @ 250V , Non-Inductive Operating Temperature: 20C to 50C Enclosure: IP65 / TYPE 4X / NEMA 4X

159134

Class I, Div.2, Gr. A, B, C & D T3C Class II, Div. 2, Gr. F & G Class III Per Drawing A5E03936871

159134

Class I, Div.2, Gr. A, B, C & D T3C Class II, Div. 2, Gr. F & G Class III Per Drawing A5E03936871

159134

Class I, Div.2, Gr. A, B, C & D T3C Class II, Div. 2, Gr. F & G Class III Per Drawing A5E03936871

Siemens Milltronics Process Instruments, Peterborough Assembled in Canada with domestic and imported parts

Siemens Milltronics Process Instruments, Peterborough Assembled in Canada with domestic and imported parts

Siemens Milltronics Process Instruments, Peterborough Assembled in Canada with domestic and imported parts

SITRANS LUT420 7MLxxxx-xxxxx-xxxx Serial No.: GYZ / B1034567 Power Rating: 10 32V , 10W Contact Rating: 1A/5A @ 250V , Non-Inductive Operating Temperature: 20C to 50C Enclosure: IP65 / TYPE 4X / NEMA 4X

SITRANS LUT430 7MLxxxx-xxxxx-xxxx Serial No.: GYZ / B1034567 Power Rating: 10 32V , 10W Contact Rating: 1A/5A @ 250V , Non-Inductive Operating Temperature: 20C to 50C Enclosure: IP65 / TYPE 4X / NEMA 4X

SITRANS LUT440 7MLxxxx-xxxxx-xxxx Serial No.: GYZ / B1034567 Power Rating: 10 32V , 10W Contact Rating: 1A/5A @ 250V , Non-Inductive Operating Temperature: 20C to 50C Enclosure: IP65 / TYPE 4X / NEMA 4X

159134

Class I, Div.2, Gr. A, B, C & D T3C Class II, Div. 2, Gr. F & G Class III Per DWG. A5E03936871

159134

Class I, Div.2, Gr. A, B, C & D T3C Class II, Div. 2, Gr. F & G Class III Per DWG. A5E03936871

159134

Class I, Div.2, Gr. A, B, C & D T3C Class II, Div. 2, Gr. F & G Class III Per DWG. A5E03936871

Siemens Milltronics Process Instruments, Peterborough Assembled in Canada with domestic and imported parts

Siemens Milltronics Process Instruments, Peterborough Assembled in Canada with domestic and imported parts

Siemens Milltronics Process Instruments, Peterborough Assembled in Canada with domestic and imported parts

CSA luokka I, jako 2 -kytkentpiirustus numero A5E03936871 voidaan ladata tuotesivulta sivustostamme osoitteessa www.siemens.com/sitransLUT400.

Rjhdysvaarallisten tilojen asennusta koskevat ohjeet (ATEX-direktiivi 94/9/EY, liite II, kohta 1/0/6)
Seuraavat ohjeet koskevat sertifikaatin SIRA 12ATEX9143X mukaisia laitteita: 1) 2) 3) 4) Kytt- ja kokoamisohjeet ovat pohjeissa. Laite on sertifioitu luokan 3D laitteeksi. Laitetta saadaan kytt tilaluokan 22 alueella. Laitteiden suurin pintalmptila on T70 C (ympristn lmptilassa +50 C). Laitteiston valinta plyn syttymislmptilan perusteella on tarkistettava sovellettavan kytnnn mukaisesti. 5) Laite on sertifioitu kytettvksi 20+50 C:n ympristn lmptilassa. 6) Kaapelien tai johdinten sisntulojen on tytettv direktiivin 94/9/EY vaatimukset, jotka koskevat ryhm II, luokkaa 3D, ja niiden on silytettv kotelon IP-luokitus. 7) Kaapelin valinnassa on noudatettava sovellettavaa kytnt, ja kaapelin eristyksen on kestettv kotelon suurin pintalmptila (T70 C). 8) Laitetta ei ole luokiteltu direktiivin 94/9/EY liitteen II kohdan 1.5 mukaiseksi turvallisuuteen mytvaikuttavaksi laitteeksi. 9) Tmn laitteen saa asentaa ja tarkistaa vain koulutettu henkil siihen sovellettavan kytnnn ja standardin (EN 60079-14 ja EN 60079-17 Euroopassa) mukaisesti. 10) Tmn laitteen saa korjata vain koulutettu henkil sovellettavan kytnnn ja standardin (esim. Euroopassa EN 60079-19) mukaisesti.

Suomi
Sivu FI-36

SITRANS LUT400 (HART)

7ML19985XU81

11) Laitteeseen liitettvt tai vaihdettavat osat saa asentaa vain koulutettu henkil valmistajan ohjeiden mukaisesti. 12) Manuaaliseen ohituskytkentn voidaan kytt erotuskytkint, joka toimitetaan kenttasennuksen mukana.

ERITYISET EHDOT TURVALLISTA KYTT VARTEN


X-pte sertifiointinumerossa tarkoittaa, ett laitteen turvalliseen kyttn sovelletaan seuraavia erikoisehtoja: Kaapelien tai suojaputkien sisntulojen on tytettv direktiivin 94/9/EY vaatimukset, jotka koskevat ryhm II, luokkaa 3D, ja niiden on silytettv kotelon IP-suojausluokka. Kaapelin valinnassa on noudatettava sovellettavaa kytnt, ja kaapelin eristyksen on kestettv kotelon suurin pintalmptila (T70 C). Kotelo ei johda shk ja saattaa synnytt tietyiss riolosuhteissa shkstaattisia varauksia, jotka voivat sytytt materiaaleja. Kyttjn tulee varmistaa, ett laitetta ei asenneta ympristn, jossa se altistuu ulkoisille olosuhteille (kuten korkeapaineiselle hyrylle), jotka saattavat synnytt shkstaattisia varauksia shk johtamattomille pinnoille. Lisksi on huomattava, ett laitteen saa puhdistaa vain kostealla kankaalla. Laite on asennettava alueelle, jolla mekaanisen vahingoittumisen vaara on pieni, tai jollei tm ole mahdollista - se on suojattava lissuojalla, joka tarjoaa tarkoitukseen soveltuvan mekaanisen suojauksen. Etasennettavalla paneeliasenteisella kannella varustettu LUT400-sarjan ohjain (LUT420, LUT430, LUT440) on asennettava suojausluokaltaan asianmukaiseen koteloon, joka tytt direktiivin 94/9/EY vaatimukset koskien ryhm II, luokkaa 3D ja jonka suojausluokka on vhintn IP65.

Suomi

7ML19985XU81

SITRANS LUT400 (HART)

Sivu FI-37

Huomautukset

Suomi
Sivu FI-38

SITRANS LUT400 (HART)

7ML19985XU81

SITRANS LUT400 (HART) Snabbstartsmanual


Denna manual ger en versikt ver de viktigaste egenskaperna och funktionerna hos SITRANS LUT400 (HART 1)). Vi rder er uttryckligen att skaffa er den detaljerade versionen av manualen s att ni kan f ut det mesta av ert instrument. De fullstndiga SITRANS LUT400 (HART) Driftsinstruktionerna, liksom SITRANS LUT400 (HART) Kommunikationsmanual, kan laddas ner frn produktsidan p vr webbplats p: www.siemens.com/sitransLUT400. Den tryckta manualen kan anskaffas frn er lokala Siemens-representant. Frgor angende innehllet i denna manual kan sndas till: Siemens AG Siemens Milltronics Process Instruments 1954 Technology Drive, P .O. Box 4225 Peterborough, Ontario, Kanada, K9J 7B1 E-post: techpubs.smpi@siemens.com Copyright Siemens AG 2012. Med ensamrtt
Vi uppmuntrar anvndare att anskaffa de auktoriserade inbundna manualerna eller studera elektroniska versioner som sammanstllts av Siemens Milltronics Process Instruments, Inc. Siemens Milltronics Process Instruments Inc. ansvarar inte fr innehllet i utdrag frn eller hela reproduktioner av bde tryckta och elektroniska versioner.

Svenska

Ansvarsbegrnsning
Trots att vi kontrollerat att innehllet i denna handbok verensstmmer med beskriven instrumentering, kan avvikelser frekomma. Drfr kan vi inte garantera full verensstmmelse. Innehllet i denna handbok revideras regelbundet och ndringar tas med i fljande versioner. Vi vlkomnar alla frslag till frbttringar. Tekniska data kan komma att ndras utan fregende varsel.

MILLTRONICS r ett registrerat varumrke fr Siemens Milltronics Process Instruments.

Teknisk support
Support ges dygnet runt. Fr att hitta ditt lokala Siemens Automation-kontors adress, telefonnummer och faxnummer, g till: www.siemens.com/automation/partner: Klicka p fliken Contact (Kontakt) fr att hitta din produktgrupp (till exempel: Automation Technology (Automationsteknologi)). Klicka igen p produktgruppens rullgardinsmeny fr att komma till produktfamilj. Fr niv: +Automation Technology > Sensor Systems (+Sensorsystem) >+Process Instrumentation (+Processinstrumentering) > +Level Measurement (+Nivmtning) > +Continuous (+Kontinuerlig). Vlj Ultrasonic (Ultraljud). Fr flde: +Automation Technology > +Sensor Systems >+Process Instrumentation > +Flow Measurement (+Fldesmtning). Vlj Ultrasonic. Vlj land och sedan Stad/Region. Vlj Technical Support (Teknisk support) under Service. Fr on-line teknisk support, g till: www.siemens.com/automation/support-request Fr in apparatnamn (SITRANS LUT400) eller ordernummer, klicka sedan p Search (Sk), och vlj motsvarande produkttyp. Klicka p Next (Nsta). Fr in ett nyckelord som beskriver din frga. Blddra igenom motsvarande dokumentation, eller klicka p Next fr att skicka en e-post med din frga till Siemens Tekniska Support-stab. Siemens IA&DT Technical Support Center (Siemens IA&DT Teknisk Support-central): telefon +49 (0)911 895 7222
1)

HARTr ett registrerat varumrke fr HART Communications Foundation.

7ML19985XU81

SITRANS LUT400 (HART)

Sida SV-1

Riktlinjer fr skerhet
Svenska
Varningsnotiser mste iakttas fr att din personliga och andras skerhet ska kunna skerstllas samt skydda produkten och ansluten utrustning. Dessa varningsnotiser tfljs av anvisningar om den skerhetsniv som skall observeras. VARNING-symbolen hnvisar till en varningssymbol p produkten. Den informerar om att underltenhet att vidta erforderliga frsiktighetstgrder kan orsaka ddsfall, svr kroppsskada och/eller stor materiell skada. Denna VARNING-symbol som anvnds nr det inte finns ngon motsvarande varningssymbol p produkten, informerar om att underltenhet att vidta erforderliga frsiktighetstgrder kan orsaka ddsfall, svr kroppsskada och/ eller stor materiell skada. Anmrkning:innebr viktig information om produkten eller den delen av anvndarmanualen.

FCC-verensstmmelse
Endast fr anlggningar i USA: Regler uppstllda av Federal Communications Commission (FCC)
VARNING: ndringar eller modifikationer som inte uttryckligen godknts av Siemens Milltronics kan upphva anvndarens rtt att bruka utrustningen. Anmrkningar: Denna utrustning har testats och befunnits klara de grnser som uppsatts fr digital utrustning av Klass A, i enlighet med Del 15 av FCC-reglerna. Dessa grnser har satts s att skligt skydd mot skadlig strning uppns nr utrustningen anvnds i industriell milj. Denna utrustning genererar, anvnder och kan utstrla radiofrekvensenergi samt kan, om den inte installeras och anvnds i enlighet med anvndarmanualen, frorsaka besvrande strningar av radiokommunikationer. Om denna utrustning anvnds i bostadsomrden uppstr frmodligen besvrande strningar p radiokommunikationerna, och anvndaren kan d lggas att avhjlpa strningarna p egen bekostnad.

Sida SV-2

SITRANS LUT400 (HART)

7ML19985XU81

SITRANS LUT400
VARNING: SITRANS LUT400 fr endast anvndas p det stt som anges i denna manual; i annat fall kan det skydd, som utrustningen erbjuder, visa sig otillrckligt.

Svenska

Anmrkning: Denna produkt r avsedd att anvndas i industriomrden. Om denna utrustning anvnds i bostadsomrden kan den frorsaka strningar p mnga frekvensbaserade kommunikationer. Siemens SITRANS LUT400serie kontrollapparater bestr av kompakta, enpunktsultraljudskontrollenheter med stor rckvidd avsedda fr kontinuerlig nivmtning av vtskor, slam och fasta kroppar, och vervakning med hg precision av ppet kanalflde. Serien r kompatibel med hela Siemens EchoMax-sortiment Transduktorer, som ger ett aktionsomrde p 0,3 till 60 meter (beroende p transduktor). SITRANS LUT400 har utrustats med ett bakifrn upplyst Local User Interface (Lokalt Anvndargrnssnitt) (LUI), som visar funktionerna fr menyprogrammering. Den r frsedd med ett intuitivt webbaserat grnssnitt, och hjlpprogram fr plug and play utfrande. SITRANS LUT400 stder HARTs kommunikationsprotokoll, och flera programvaror fr fjrrkommunikationer [SIMATIC PDM, AMS Device Manager, Field Communicator 375/475, och Field Device Tools (FDTs)]. Den visar ven vr kommande genaration av Sonic Intelligence, som ytterligare kommer att frstrka vr ledande mtkapacitet inom industrin, allt medan den blir enklare att anvnda.

Specifikationer
Effektfrbrukning

Fr fullstndig lista, var god se SITRANS LUT400 (HART) Driftsinstruktioner.

Vxelstrmsversion

100-230 V vxelstrm 15%, 50 / 60 Hz, 36 VA (10W)1) Skring: 5 x 20 mm, Trg, 0,25A, 250V
Angiven effektfrbrukning r den maximala.

Likstrmsversion
10-32 V likstrm, 10W1 Skring: 5 x 20 mm, Trg, 1,6A, 125V

1)

Miljvillkor
Hjd ver hav
2 000 m max. 4

Omgivande temperatur
-20 till +50 C (-4 till +122 F)

Installationskategori
II

Nedsmutsningsgrad Relativ fuktighet


Passar utomhus (endast med IP65 / Typ 4X / NEMA 4X Kapsling)

Godknnanden
Anmrkningar: Apparatens mrkskylt listar de godknnanden som gller fr er apparat. Anvnd lmpliga ledningsttningar fr att upprtthlla IP- eller NEMA-klassning. Allmnt Riskfylld Icke antndbar (Kanada) Dammantndninggssker (Internationellt/Europa) CSAUS/C, CE, FM, UL-listad, C-TICK CSA Klass I, Div. 2, Grupp A, B, C, D; Klass II, Div. 2, Grupp F, G; Klass III ATEX II 3D Ex tc IIIC T70 C IP65 IECEx SIR 12.0057X SIRA 12ATEX9143X SITRANS LUT400 (HART) Sida SV-3

7ML19985XU81

Installation
Svenska
Anmrkningar: Installation fr endast utfras av kompetent personal och enligt gllande lokala bestmmelser. Produkten kan ge upphov till elektrostatiska sttar. Flj lmpliga jordningsrutiner. All kabeldragning mste ha isolering lmpad fr minst 250V. Farlig spnning ligger p transduktorterminaler under drift. DC-ingngarna skall komma frn en klla som ger elektrisk isolering mellan ingng och utgng, fr att uppfylla tillmpliga skerhetskrav i IEC 61010-1 Relkontaktterminaler anvnds med utrustning som inte har ngra tillgngliga spnningsfrande delar och kabling med isolering fr minst 250 V. Maximal tillten driftspnning mellan nrliggande relkontakter r 250 V. Det icke-metalliska hljet ger inte jordning mellan ledningsanslutningar. Anvnd jordade bussningar och byglingar.

Monteringsanvisningar
Monteringsanvisningarna skiljer sig t fr vgg, rr, DIN-skena, och monteringsanordningar fr fjrrdisplay. Flj de speciella anvisningarna fr din apparat.

Anmrkning: Nr du utfr en kabelgenomfrning, flj anvisningarna fr Kabelgenomfrning p sida 9 innan du monterar SITRANS LUT400. Montering p vgg eller panel
Alla konfigurationer av SITRANS LUT400 transporteras med en bakre platta fr montering. Modellen fr montering p panel levereras bde med en LUI-skrm och med en tom lucka.
bakre monteringsplatta

kapsling med skrm p luckan som tillval

klmmor (2 platser) klmmor (2 platser)

Anmrkning: Fstanordningar fr vggmontering medfljer ej.

Sida SV-4

SITRANS LUT400 (HART)

7ML19985XU81

bakre monteringsplatta

Svenska

kapsling med tom lucka

klmmor (2 platser) klmmor (2 platser)

Anmrkning: Fstanordningar fr vggmontering medfljer ej. kapslingens baksida

uppslitsade detaljer

Vggmontering av kapslingen
1) 2) 3) Mrk ut och borra upp fyra hl i monteringsytan fr de fyra skruvarna (tillhandahlls av kunden). Skruva i dem med en skruvmejsel. Mrk ut de uppslitsade detaljerna p apparatens baksida med klmmor p bakre monteringsplatta. Tryck LUT400 jmt mot den bakre plattan och lt den glida nert s att den stter sig p plats.

skruvhl fr montering p den bakre plattan

Observera:
Rekommenderad storlek fr fstskruv: M8 eller 5/16" skruv med bricka p maximalt 17 mm eller 5/8" diameter utvndigt Rekommenderad montering: montera direkt p vgg. Om alternativ monteringsyta vljs, SKA den kunna bra fyra gnger apparatens vikt.

7ML19985XU81

SITRANS LUT400 (HART)

Sida SV-5

Fjrrmonterad lucka

Svenska

fjrrdisplaylucka

packning

4 fstanordningar fr panelmontering medfljer frlngningskabel SITRANS LUT400

Montering av fjrrdisplay
Anmrkning: Fjrrdisplay kan monteras upp till 5 m frn apparaten med hjlp av tv frlngningskablar som tillbehr (vardera 2,5 m lng). Se fullstndiga Driftsinstruktioner. 1) Tag med den medfljande mallen ut det ndvndiga hlet fr den fjrrmonterade LUIskrmen. Placera packningen innanfr luckan, s att monteringshlen hamnar p rad. Placera baksidan av fjrrdisplayen mot i panelen urtagna hl. Mrk ut och borra upp fyra hl i monteringsytan fr de fyra skruvarna (medfljande). Fst med en skruvmejsel och drag t. Anmrkning: Rekommenderat vridmoment vid faststtning av skruvar fr god ttning: 1,1 N m 10 in-lbs

2)

Observera:
Rekommenderad montering: montera p panel, upp till 5 m frn apparaten. Om alternativ monteringsyta anvnds, SKA denna kunna bra fyra gnger apparatens vikt. Anmrkning: Medfljande fstanordningar: M5 skruv, ttningsbricka, M5 platt bricka och mutter. Dessa anordningar fordras fr att upprtthlla IP65-klassning p fjrrmonterad lucka. Sida SV-6 SITRANS LUT400 (HART) 7ML19985XU81

Rrmontering

Svenska

montering av bakre platta sadelkrampa kapsling med display lucka som tillval

u-bult

rr Anmrkning: Fstanordningar fr montering av rr levereras ej.

Montering av kapslingen
1) 2) 3) Fst den bakre monteringsplattan p rret med hjlp av till rrets diameter avpassade ubultar och sadelkrampor, (tillhandahlls av kunden). Fst bultarna genom att dra t. Drag inte t fr hrt s att plattan vrids eller bjs. I s fall kanske LUT400 inte fster mot den bakre plattan. Fst apparaten mot den bakre monteringsplattan (s som beskrivs i steg 3 i Vggmontering av kapslingen p sida 5).

Observera:
Rekommenderad montering: direkt mot horisontellt eller vertikalt rr. Om alternativ monteringsyta anvnds, SKA denna kunna bra fyra gnger apparatens vikt. Rekommenderade rrdimensioner: maximum: 3" rr, minimum: 3/4" rr Rekommenderade storlekar fr fstningsanordningar: U-Bultar: maximum: 3" rrstorlek med M8 eller 3/8" gnga minimum: 3/4" rrstorlek med M6 eller 1/4" gnga Sexkantsmuttrar: M6 eller 1/4" till M8 eller 3/8" Brickor: maximum: 16 mm eller 13/16" diameter utvndigt.

7ML19985XU81

SITRANS LUT400 (HART)

Sida SV-7

Montering DIN-skena
kapslingens baksida

Svenska

DIN-skena

klmmor (2)

urtag

Glidbana DIN-skena

Montering av kapslingen
1) 2) 3) Kapslingens hrnvinkel mot DIN-skena, placerad strax ver skenans topp. Fr kapslingen nert mot DIN-skenan fr att haka klmmorna p kapslingens baksida p toppen av DIN-skenan. Tryck apparaten ttt intill DIN-skenan fr att fra in DIN-skenans glidbana, s att kapslingen fster skert mot DIN-skenan.

Observera:
Rekommenderad montering: direkt mot horisontell DIN-skena. Erfordrade dimensioner fr DIN-skenan: TH 35-7,5 eller TH 35-15 enligt standard IEC 60715. DIN-skenan SKA kunna bra fyra gnger SITRANS LUT400:s vikt.

Borttagning ev kapsling
1) 2) Placera en skruvmejsel frn apparatens framsida i en springa nederst p DIN-skenans glidbana och bnd nert. D lossnar glidbanan nertill p DIN-skenan. Fortstt att hlla ner glidbanan, medan du fr kapslingen uppt fr att lossa klmmorna verst p DIN-skenan.

Sida SV-8

SITRANS LUT400 (HART)

7ML19985XU81

Frberedelse fr kabelgenomfrning
Kablarna kan fras genom rr eller fras in i kapslingen genom packboxar. Flj steg 1 till 5 nedan fr att frst frilgga kabelingngshlen, komplettera sedan med stegen fr anvndning av rr, eller packboxar. 1) Frskra dig om att kapslingsluckan r stngd och skruvarna tdragna. 2) Placera spetsen av en skruvmejsel i skran p den utdragbara flikens yttre diameter (se fljande bild). 3) Sl p skruvmejseln med handflatan fr att frilgga ingngshlet. 4) Lossa skruvarna och tag bort kapslingsluckan. 5) Tag bort plastdelarna som tcker kapslingens ingngshl. Se till att inte skada elektroniska delar med statisk elektricitet, eller verktygen du anvnder fr att frilgga ingngshlen.
SITRANS LUT400

Svenska

ingngshl fr kablar

utdragbara flikar (3 platser)

skra fr skruvmejsel (3 platser)

Anmrkningar: Nr kabelingngshlens utdragbara flikar tagits bort, r ingngshlet 21,4 mm till 21,6 mm i diameter. M20 packbox (20 mm i diameter), och 1/2" NPT rr (21,3 mm i diameter) passar till detta ingngshl. Var frsiktig nr du vljer lmplig ttning av ingngshlen. Plan packning rekommenderas (i stllet fr Oring). Om alternativa packboxar anvnds, ansvarar kunden fr att IP65-klassningen vidhlls fr ingngshlen.

Kabel frd genom rr


(fortsttning frn steg 1 till 5 ovan) 6) Efter att ha frberett kabelns infrning frn steg 1 till 5 ovan, fst ledningen mot kapslingen med hjlp av beslag av lmplig storlek godknt fr vattentta tillmpningar. (Rrstorleken r 1/2" NPT.) 7) Stt tillbaka kapslingens lucka och dra t skruvarna.

Oskyddad kabel och infrning genom packboxar


(fortsttning frn steg 1 till 5 ovan) 6) Efter att ha frberett kabelinfringen enligt steg 1 till 5 ovan, lossa packboxarna och fst dem ordentligt vid kapslingen. 7) Tr kablarna genom packboxarna. Se till att starkstrmskabeln hlls tskild frn signalkablarna och dra sedan kablarna till anslutningsplinterna. 8) Dra t packboxen s att du fr god ttning. 9) Stt tillbaka kapslingens lucka och dra t skruvarna.

7ML19985XU81

SITRANS LUT400 (HART)

Sida SV-9

Anslutningar
Skerhetsanmrkningar fr anslutning
Fr fullstndiga kabeldragningsinstruktioner hnvisas till de fullstndiga Driftsinstruktionerna. Kontrollera att alla systemkomponenter har installerats i enlighet med instruktionerna. Anslut alla kabelskrmningar till LUT400:s skrmade terminaler (p apparaterna betecknade med symbol ). Fr att undvika potentialskillnader i jordningen, se till att kabelskrmningarna r korrekt jordade. Hll exponerade ledare p skrmade kablar s korta som mjligt fr att minska bruset p linjen frorsakat av krypstrmmar och ljudupptagning.

Svenska

Anslutning SITRANS LUT400


VARNINGAR:

Se mrkskylt p instrumentet, fr att kontrollera godknnandeklassningen. Anvnd lmpliga ledningsttningar fr att upprtthlla IP- eller NEMAklassning.
Anmrkningar: Separata kablar och ledningar kan krvas fr att uppfylla standardiserad kabeldragningspraxis, eller elbestmmelser. Fr att komma till kablingsdelen: 1) 2) 3) Lossa lsskruvarna ett 1/4 varv. Lyft luckan uppt och till vnster p dess gngjrn. Luckan kan frbli ppen p sina gngjrn, eller kan den lyftas ur gngjrnen och stllas t sidan fr att ge tillgng till kablingsdelen.

4) 5) 6)

Utfr alla anslutningar enligt fljande instruktioner. Nr kablingen r klar, stt tillbaka apparatens lucka. Drag t lsskruvarna.

Sida SV-10

SITRANS LUT400 (HART)

7ML19985XU81

Kablingsdel
Plinten p LUT400 har plats fr anslutning av alla in- och utgngar samtidigt. Ledare kan tas bort fr att underltta kablingen.

Svenska

OBS: Se till att ledarna hamnar rtt under terinstallationen. I annat fall kan skada uppst p apparaten, eller drtill hrande extern utrustning. Anmrkning: Rekommenderat vridmoment p anslutningsklmskruvar. 0,56 - 0,79 N m 5 - 7 in-lbs Drag inte t skruvarna fr hrt Strm
TS-3 TRANSDUCER DISCRETE INPUTS INPUT POWER

strmfrsrjning

8 1

9 2

10

11 0.25 A, 250 V SLOW BLOW

12 L1

13 L2/N 100 - 230 V

14

DC SYNC

SHLD

WHT

BLK

BLK

WHT

plint
mA OUT/HART BATTERY BR2032

batteri

RELAYS

DISPLAY 15 16 RLY 1 17 18 19 20 21 22 23 24

RLY 2

RLY 3 ACT

PAS SERVICE

USB anslutning

displaykabel

mA HART anslutning

Effekt
Varningar: DC-ingngarna skall komma frn en klla som ger elektrisk isolering mellan ingng och utgng, fr att uppfylla tillmpliga skerhetskrav i IEC 61010-1 Allt fltkablage mste ha isolering lmpad fr mrkspnningarna. Viktigt! Innan du kopplar in strm p SITRANS LUT400 fr frsta gngen, frskra dig om att all inkopplad larm-/styrutrustning r frnkopplad till dess att tillfredsstllande systemdrift och prestanda konstaterats.

7ML19985XU81

SITRANS LUT400 (HART)

Sida SV-11

SITRANS LUT400 kan fs i modell fr vxel- eller likstrm.

Vxelstrm Svenska
INPUT POWER

Likstrm
INPUT POWER

12

13

14

12 10 - 32 V

13

14

Vxelstrm: 100-230 V vxelstrm 15%, 50/60 Hz, 36 VA (10W)

L1 L2/N 100 - 230 V

Likstrm: 10-32 V likstrm, 10W

Anmrkning: Frskra dig om att apparaten r ansluten till tillfrlitlig jordning. 1) 2) 3) Fr strmanslutning strippa kabelisoleringen ungefr 70 mm (2.75") frn nden p kabeln och tr ledningarna genom packboxen 1). Anslut ledningarna till uttagen som visas: polariteten finns angiven p anslutningsplinten. Jorda apparaten i enlighet med lokala bestmmelser. Anm. fr Vxelstrmanslutningar Utrustningen mste skyddas med en 15 A skring eller ett verspnningsskydd i byggnaden. Ett verspnningsskydd eller en strmbrytare i byggnaden, mrkt som frnkopplingsbrytare, mste finnas uppsatt i omedelbar nrhet till utrustningen och inom rckhll fr operatren.

Batteriet
SITRANS LUT400 levereras med ett installerat batteri. (Se bild p sida 11 fr batteriutrymme.) Batteri (BR2032) har en livslngd p tio r, och pverkas av omgivande temperatur. Om LUT400 frlorar ingngsstrm, upprtthller batteriet driften fr apparatens realtidklocka tills strmmen kommer tillbaka. (Se de fullstndiga Driftsinstruktionerna fr byte av batteri.)

1)

Om kabel dras genom skyddsrr, anvnd endast lmpligt dimensionerade nav fr vattentta tillmpningar.

Sida SV-12

SITRANS LUT400 (HART)

7ML19985XU81

Transduktorer
Varning: Farlig spnning ligger p transduktorplintar under drift. Fr in transduktorkabeln i en jordad metalledning, separat frn allt annat kablage (utom kablar fr temperatursensor TS-3, i tillmpliga fall). Anmrkningar: Anvnd inte koaxial kabel med SITRANS LUT400 p grund av den hga spnning som verfrs till skrmningen av den koaxiala kabeln. Anslut inte LUT400- skrmning och vita transduktorkablar tillsammans, dra dem till skilda uttag. Fst inte avseende vid ldre transduktormanualer som rekommenderar sdan praxis.
TS-3 TRANSDUCER

Svenska

SHLD

BLK

WHT

Reler
Relkontakter visas i strmlst lge. Alla reler kan konfigureras som positiv eller negativ logik (se parameter 2.8.11.Relay Logic i de fullstndiga Driftsinstruktionerna).
RELAYS

BLK

WHT

Strmavbrott
Reler 2, 3 r normalt ppna. Rel 1 kan kablas antingen normalt ppen eller normalt stngd. I fall av frlust av ingngsstrm, slr relerna om till normalt status.

15

16 RLY 1

17

18

19

20

21

RLY 2

RLY 3

Relklassning
en Form C (NO eller NC) rel (rel 1), 1A vid 250 V vxelstrm, induktionsfri, 3A vid 30 V likstrm tv Form A (NO) reler (reler 2,3), 5A vid 250 V vxelstrm, induktionsfri, 3A vid 30 V likstrm

7ML19985XU81

SITRANS LUT400 (HART)

Sida SV-13

Anslutning av HART mA Uteffekt (HART) Svenska


mA OUT/HART

Fr AKTIV HART anslutning (med hjlp av LUT400 integral strmtillfrsel), anslut terminalerna 22 och 23.

22

23

24

Fr PASSIV HART anslutning (med hjlp av extern strmtillfrsel), anslut terminalerna 23 och 24.

ACT

PAS

Fr mer information hnvisas till mA utparametrar (2.5.Current Output) i de fullstndiga Driftsinstruktionernas avsnitt fr parameterreferenser.

Anslutning i installationer i riskfyllt omrde


Se sida 35

Idrifttagning SITRANS LUT400


Vi rekommenderar att du konfigurerar din tillmpning p fljande stt: Brja med att kra fr din tillmpning lmplig Snabbstartsguide (Niv, Volym, Flde). Stll sedan upp pumpar via hjlpprogrammet fr Pumpstyrning (om tillmpligt). Konfigurera till slut larm och andra styrenheter, totalisatorer och samplare, med referens till respektive parametrar (se fullstndiga Driftsinstruktioner). Det finns tv stt att f tillgng till Snabbstartsguiden: lokalt (se Snabbstartsguide via LUI p sida 17) p distans (se Andra Snabbstartsguider: p sida 17)

Aktivering av SITRANS LUT400


Inkoppling av apparaten p elntet. SITRANS LUT400 startar automatiskt i Mtlge. En inledningsbild visar frst Siemens logga och sedan visas versionen av aktuell firmware medan den frsta mtningen bearbetas. Den frsta gngen som apparaten konfigureras, uppmanas du att vlja sprk (Engelska, Tyska, Franska, Spanska eller Kinesiska). Korrekt datum och klockslag ska stllas in innan apparaten konfigureras. (Fr att gra ndringar, g till parameter 2.14.Date and Time. Se fullstndiga Driftsinstruktioner fr fler detaljer.)

Sida SV-14

SITRANS LUT400 (HART)

7ML19985XU81

LCD-display Display mtlge: Normal driftsniv 1) Niv


1 10
SITRANS LUT400 LEVEL

2
1/2

4 5

Svenska

18.91
1

6
2

5 SAVING PARAMETERS

Flde
1 10 11 12
SITRANS LUT400 FLOW HEAD
1 2 3

3
1/2
L/S M

4 5 6

500313..
60.000

DAILY TOTAL,...

999999.99

5 SAVING PARAMETERS

1 tryck in 2 uppmtt vrde (niv, utrymme, avstnd, volym, flde eller toppniv) 3 visat vrde [Primr Variabel (PV)=1 av 2, Sekundr Variabel (SV)=2 av 2] 4 toggle indikator1) fr PV eller SV 5 enheter 6 stapeldiagram indikerar niv 7 sekundrt omrde indikerar konfigurerade reler (vnster) och diskreta ineffekter (hger) 8 textytan visar statusmeddelanden 9 apparatstatusindikator 10 valt (primrt) sensorlge: niv, utrymme, avstnd, volym, vtskehjd eller flde 11 sekundrt sensorlge = topp vid primrt sensorlge = flde 12 vrden totalisator: visar vxlingar mellan daglig totalisator och totalisator i drift

Fel freligger
0 LOE

8 textyta visar en felkod och ett felmeddelande 9 service krvs-ikon visas

Visning Programlge Menyruta Navigation


En synlig menystapel indikerar att parameterTRANSDUCER 2.1.6 aktuell nummer menylistan r fr lng fr att kunna meny / NO TRANSDUCER parameterXRS-5 visa alla positioner. namn XPS-10 positionsband Positionsbandets djup i menystapeln XPS-15 aktuell XCT-8 position indikerar menylistans lngd: ett XCT-12 XPS-30 djupare band indikerar frre menystapel positioner. Positionsbandets lge indikerar den aktuella positionens ungefrliga lge i listan. Ett band halvvgs ner i menystapeln indikerar att den aktuella positionen r halvvgs ner i listan. Ett djupare band indikerar frre positioner.

1)

Tryck in

eller

fr att koppla.

7ML19985XU81

SITRANS LUT400 (HART)

Sida SV-15

Menyruta Parametrar Svenska


meny i vilken parametern befinner sig

Menyruta Redigering
2.1.6
parameternummer

SENSOR
UNITS SENSOR MODE SENSOR MODE SECONDARY

TRANSDUCER
NO TRANSDUCER XPS-10 XPS-15 XCT-8 XCT-12 XPS-30

2.1.6

XRS-5

TRANSDUCER

parameternamn

FREQUENCY

XRS-5

parametervrde/val

PROGRAM-lge
Tryck p apparatens knappar genom menyn med hjlp av , fr att komma till Programlge och ppna menyniv 1. Rulla , , .

Val av ett listat alternativ


a) Navigera till den nskade parametern. b) Tryck p HGER pil upplyst. c) Rulla till ett nytt val. d) Tryck in HGER pil fr att acceptera det. LCD:n tergr till menyruta parametrar och visar det nya valet. fr att ppna funktionen Redigera. Det aktuella valet r

ndring av ett numeriskt vrde


a) Navigera till den nskade parametern, till exempel, Lg Kalibreringspunkt (2.2.1.). b) Nr den r vald, visas det aktuella vrdet. c) Tryck p HGER pil fr att ppna funktionen Redigera: Det aktuella vrdet r upplyst. d) Anvnd VNSTER och HGER pil fr att flytta markren till den digitala position du nskar ndra. e) D alla siffror r belysta (valda), anvnd UPPT eller minska respektive siffra. f) Nr decimalpunkt r vald, anvnd UPPT decimalens lge. och NERT pil fr att ka
LOW CALIB. PT. 2.2.1

ESC

+ _ _ _ _ _ _ 2. 0 M

OK

och NERT pil

fr att ndra tills

g) Fr lmna utan att spara dina ndringar, tryck kontinuerligt p VNSTER pil

ESC lyses upp. Tryck p VNSTER pil igen fr att lmna utan att spara ndringarna. I annat fall, om det nya parametervrdet r korrekt, tryck kontinuerligt p HGER pil tills OK lyser.

h) Tryck p HGER pil fr att acceptera det nya vrdet. LCD:n tergr till menyruta parametrar och visar det nya valet. Kontrollera att det r exakt.

Sida SV-16

SITRANS LUT400 (HART)

7ML19985XU81

Snabbstartsguide
Fr att konfigurera SITRANS LUT400 fr tillmpningar som niv, volym (standardkrlform), eller flde, anvnd Snabbstartsguide via LUIsom beskrivs nedan. Ett hjlpprogram fr tillmpningar som anvnder mer komplexa krlformer finns tillgngligt via SIMATIC PDM. Se Snabbstartsguiden via SIMATIC PDM p sida 33.

Svenska

Andra Snabbstartsguider:
Andra Snabbstartsguider som anvnder andra programvaror, finns ven tillgngliga 1): SIMATIC PDM (HART) AMS (HART) FC375/475 (HART) FDTs (HART) Tabeller fr parameterkonfiguration (fr att registrera data innan apparaten konfigureras med Snabbstartsguide) r tillgngliga p vr webbplats. G till www.siemens.com/sitransLUT400 > Support > Application Guides (Tillmpningsguider).

Snabbstartsguide via LUI


1) 2) Tryck p fr att komma till Programlge. Vlj Hjlpprogram (1.), Snabbstart (1.1.), och sedan lmplig Snabbstart: Snabbstart Niv (1.1.1.), Snabbstart Volym (1.1.2.), eller Snabbstart Flde (1.1.3.). [Snabbstart Flde hjlpprogram visas p LUI fr LUT430 (Pump och Flde), och LUT440 (OCM) endast konfigurerade modeller.] Flj stegen och vlj sedan Finish (Avsluta) fr att spara ndringarna i Snabbstartsparametern och g tillbaka till Programmenyn, tryck sedan p fr att g tillbaka till Mtlget. tre gnger

3)

Anmrkningar: Snabbstartsguidens instllningar hnger samman och ndringar tillmpas frst d du anvnder Finish i slutsteget. Anvnd inte Snabbstartsguiden fr att ndra individuella parametrar: se i stllet avsnittet Parameter i de fullstndiga Driftsinstruktionerna. (Utfr kundanpassning fr din tillmpning frst efter det att Snabbstarten fullbordats.) Fr mer information angende hjlpprogram fr parameter, se de fullstndiga Driftsinstruktionerna.

1. Hjlpprogram 1.1. Snabbstart 1.1.1. Snabbstart Niv


Anvnd detta hjlpprogramfr att konfigurera enkla nivtillmpningar. a) Tryck p HGER pil fr att aktivera PROGRAM-lge och ppna menyniv 1: MAIN MENU (HUVUDMENY). b) Tryck in HGER pil c) Tryck p HGER pil
1)

fr att navigera till menyposition 1.1.1. fr att ppna Snabbstart niv.

Se Kommunikationer fr SITRANS LUT400 (HART) Manual

7ML19985XU81

SITRANS LUT400 (HART)

Sida SV-17

d) Vid varje steg, tryck p NERT pil fr att acceptera standardvrden och g direkt till nsta position,

QUICK START LEVEL


TRANSDUCER XRS-5
BACK CANCEL NEXT EDIT

Svenska

eller HGER pil fr att ppna funktionen Redigera: det aktuella valet r upplyst. e) Rulla till den nskade positionen och tryck p HGER pil fr att spara ndringen, och fr att fortstta. tryck sedan p NERT pil f)

QUICK START LEVEL


TRANSDUCER

NO TRANSDUCER XPS-10 XPS-15 XCT-8 XCT-12

XRS-5

Du kan nr som helst trycka p UPPT pil pil fr att annullera hjlpprogrammet. Start av Snabbstart Nivhjlpprogram

fr att g tillbaka, eller p VNSTER

Visar den typ av hjlpprogram som ska kras.

Alternativ
Transduktor

CANCEL (ANNULLERA), START (STARTA)

Specifierar den Siemens transduktor som r ansluten till apparaten.

Alternativ
Drift

NO TRANSDUCER (INGEN TRANSDUKTOR), XRS-5, XPS-10, XPS-15, XCT-8, XCT-12, XPS-30, XPS-40, XLT-30, XLT-60, STH Standard: NO TRANSDUCER

Reglerar den typ av mtning (och motsvarande mA uteffekt) som erfordras av tillmpningen.
sensornollpunkt vre kalibreringspunkt volym

avstnd

niv Undre kalibreringspunkt

Avlgsna Mtomrdet

Lge
LEVEL (NIV) SPACE (VOLYM) DISTANCE (AVSTND)

Beskrivning
* Materialhjd

Referenspunkt
Undre kalibreringspunkt (process tom niv) vre kalibreringspunkt (process full niv) sensornollpunkt (forts. p nsta sida)

Avstnd till materialyta

Sida SV-18

SITRANS LUT400 (HART)

7ML19985XU81

(fortsttning frn fregende sida)

Lge

Beskrivning
Vlj INTE. Om Driftsvrdet visas som ANNAT, har apparaten konfigurerats som nivkontrollenhet, men har tidigare stllts in i annat lge n NIV, VOLYM eller AVSTND. Driftsvrdet mste regleras som NIV, VOLYM eller AVSTND fr att kras med Snabbstart Niv hjlpprogram.

Referenspunkt

Svenska

ANNAT

Temperaturklla

Avlsning av temperaturklla anvnd fr justering av ljudets hastighet.

Alternativ

TRANSDUCER (TRANSDUKTOR), FIXED TEMPERATURE (FASTSTLLD TEMPERATUR), EXTERNAL TS-3 (EXTERN TS-3), AVERAGE OF SENSORS (GENOMSNITT SENSORER) Standard: TRANSDUCER

Faststlld Temperatur

Anvnd detta tecken om ingen temperatursensorapparat anvnds.

Vrde

Omrde: -100,0 till +150,0 C Standard: +20,0 C

Denna parameter visas endast om FIXED TEMPERATURE (FASTSTLLD TEMPERATUR) valts fr Temperaturklla. Enheter

Sensormtenheter.

Alternativ

M, CM, MM, FT, IN Standard: M

Anmrkning: I detta exempel, antas alla vrden vara i meter (m).


vre kalibreringspunkt

Avstnd frn Sensornollpunkt till vre kalibreringspunkt: vanligen process full niv..

Vrde

Omrde: 0,000 till 60,000 Standard: 0,000

Undre kalibreringspunkt

Avstnd frn Sensornollpunkt till Undre kalibreringspunkt: vanligen process tom niv..

Vrde

Omrde: 0,000 till 60,000 Standard: 60,000

7ML19985XU81

SITRANS LUT400 (HART)

Sida SV-19

Svarshastighet

Stller in enhetens reaktionshastighet p mtvrdesndringar i mlomrdet.

Anmrkningar: Svenska

Svarshastigheten kan endast stlls in med Snabbstartsguiden, och alla ndringar som grs p Fill Rate per Minute (2.3.1.), Empty Rate per Minute (2.3.2.), eller Damping Filter (2.3.3.) parametrarna efter fullfljandet av hjlpprogrammet kommer att trnga undan instllningarna fr Svarshastighet. Svarshastighet visas alltid som m/minut.
SLOW (LNGSAM) (0,1 M/MIN) MEDIUM (1,0 M/MIN) FAST (SNABB) (10 M/MIN) Standard: SLOW (0,1 M/MIN)

Alternativ

Anvnd en instllning just ver den maximala fyllnings- eller tmningshastigheten (den som r strst av dem). Lngsammare instllningar ger hgre precision; snabbare instllningar ger strre nivvariationer. Slut p Snabbstart nivhjlpprogram

Fr att lyckas med Snabbstart, mste alla ndringar tillmpas.

Alternativ

BACK (TILLBAKA), CANCEL (ANNULERA), FINISH (AVSLUTA) (Displayen gr tillbaka till 1.1 Snabbstartsguidens meny nr Snabbstart har fullfljts med framgng eller annullerats. Om CANCEL vljs, frs inga ndringar in i apparaten.)

Fr att verfra Snabbstartsvrdena till apparaten och terg till Programmenyn, tryck p NERT pil (Finish). Tryck sedan p VNSTER pil komma tillbaka till Mtlget. tre gnger fr att

1.1.2. Snabbstart Volym


Anvnd detta hjlpprogram fr att konfigurera volymtillmpningar med hjlp av standardkrlformer. a) Tryck in HGER pil MENU. b) Tryck in HGER pil c) Tryck p NERT pil fr att aktivera PROGRAM-lge och ppna menyniv 1: MAIN tv gnger fr att navigera till menyposition 1.1.1. och HGER pil fr att ppna Snabbstart Volym.
QUICK START VOLUME
TRANSDUCER XRS-5
BACK CANCEL NEXT EDIT

d) Vid varje steg, tryck p NERT pil fr att acceptera standardvrden och flytta direkt till nsta position, eller tryck p HGER pil fr att ppna funktionen Redigera: det aktuella valet r upplyst. e) Rulla till den nskade positionen och tryck p HGER pil f) fr att spara ndringen, och tryck sedan p fr att fortstta. fr att g fr att annullera NERT pil

QUICK START VOLUME


TRANSDUCER

NO TRANSDUCER XPS-10 XPS-15 XCT-8 XCT-12

XRS-5

Du kan nr som helst trycka p UPPT pil tillbaka, eller VNSTER pil hjlpprogrammet.

Sida SV-20

SITRANS LUT400 (HART)

7ML19985XU81

Start av Snabbstartens Volymhjlpprogram

Visar den typ av hjlpprogram som ska kras.

Alternativ
Transduktor

CANCEL (ANNULLERA), START (STARTA)

Svenska

Specifierar vilken Siemens transduktor som r ansluten till apparaten.

Alternativ
Temperaturklla

NO TRANSDUCER (INGEN TRANSDUKTOR), XRS-5, XPS-10, XPS-15, XCT-8, XCT-12, XPS-30, XPS-40, XLT-30, XLT-60, STH Standard: NO TRANSDUCER

Avlsning av temperaturklla anvnds fr att justera ljudets hastighet.

Alternativ

TRANSDUCER (TRANSDUKTOR), FIXED TEMPERATURE (FASTSTLLD TEMPERATUR), EXTERNAL TS-3 (EXTERN TS-3), AVGERAGE OF SENSORS (GENOMSNITT SENSORER) Standard: TRANSDUCER

Faststlld Temperatur

Anvnd detta knnetecken om ingen temperaturavknningsapparat anvnds.

Vrde

Omrde: -100,0 till +150,0 C Standard: +20,0 C

Denna parameter visas endast om FIXED TEMPERATURE valts som Temperaturklla. Krlform

Definierar krlformen och gr s att SITRANS LUT400 berknar volym i stllet fr niv. Om NONE (INGEN) har valts, utfrs inte ngon volymkonversion. Vlj den krlform som motsvarar det vervakade krlet eller behllaren.
NONE (INGEN), LINEAR (LINJR), CYLINDER (CYLINDRISK), PARABOLIC BOTTOM (PARABOLISK BOTTEN), HALF SPHERE BOTTOM (HALVSFRISK BOTTEN), FLAT SLOPED BOTTOM (FLAT LUTANDE BOTTEN), PARABOLIC ENDS (PARABOLISKA NDAR), SPHERE (SFRISK), CONICAL BOTTOM (KONISK BOTTEN), CURVE TABLE (KURVAT PLAN), LINEAR TABLE (LINJRT PLAN) Standard: LINEAR

Alternativ

Om CURVE TABLE eller LINEAR TABLE valts, fr in vrdena fr niv- och volymbrytpunkterna efter att ha fullfljt hjlpprogrammet. (Se parameter 2.6.7.Table 18 i de fullstndiga Driftsinstruktionerna.) Enheter

Sensormtenheter.

Alternativ

M, CM, MM, FT, IN Standard: M

Anmrkning: I detta exempel, antas alla vrden vara i meter (m).

7ML19985XU81

SITRANS LUT400 (HART)

Sida SV-21

vre kalibreringspunkt

Avstnd frn Sensornollpunkt till vre kalibreringspunkt: vanligen process full niv.

Svenska

Vrde

Omrde: 0,000 till 60,000 Standard: 0,000

Undre kalibreringspunkt

Avstnd frn Sensornollpunkt till Undre kalibreringspunkt: vanligen process tom niv.

Vrde
Svarshastighet

Omrde: 0,000 till 60,000 Standard: 60,000

Stller in enhetens reaktionshastighet p mtvrdesndringar i mlomrdet.

Anmrkningar:

Svarshastigheten kan endast stllas in med Snabbstartsguiden, och alla ndringar som grs i Fill Rate per Minute (2.3.1.) eller Empty Rate per Minute (2.3.2.) parametrarna efter fullfljandet av hjlprogrammet kommer att trnga undan reglaget Svarshastighet. Svarshastighet visas alltid i m/minut.
SLOW (0,1 M/MIN) MEDIUM (1,0 M/MIN) FAST (10 M/MIN) Standard: SLOW (0,1 M/MIN)

Alternativ

Anvnd en instllning just ver den maximala fyllnings- eller tmningshastigheten (den som r strst av dem). Lngsammare instllningar ger hgre precision; snabbare instllningar ger strre nivvariationer. Dimension A

Hjden av krlets botten d den r konisk, pyramidal, parabolisk, sfrisk, eller flat lutande..

Vrde
Dimension L

Omrde: 0,000 till 99,999 Standard: 0,000

Lngden av den cylindriska sektionen av en horisontell parabolisk krlnde.

Vrde
Volymenheter

Omrde: 0,000 till 99,999 Standard: 0,000

Avgr volymmtenheterna.

Alternativ

L, USGAL, IMPGAL, CUM, USER DEFINED (ANVNDARDEFINIERAD) * Standard: L

* Om USER DEFINED alternativ valts, mste vrdet stllas in efter det att hjlpprogrammet fullfljts. Se parameter 2.6.6.User Defined Unit i de fullstndiga Driftsinstruktionerna.

Sida SV-22

SITRANS LUT400 (HART)

7ML19985XU81

Maximal Volym

Krlets maximala volym. Fr in den krlvolym som motsvarar den Hgre Kalibreringspunkten. Om din maximala krlvolym till exempel r 8000 L, fr in vrdet 8000.

Svenska

Vrde

Omrde: 0,0 till 9999999 Standard: 100,0

Slut p Snabbstartens Volymhjlpprogram

Fr att Snabbstarten ska lyckas, mste alla ndringar tillmpas.


BACK, CANCEL, FINISH (Visning tergr till 1.1 Snabbstartsguidens meny nr

Alternativ Snabbstarten r genomfrd med framgng eller annullerad. Om CANCEL vljs,


frs inga ndringar in i apparaten.)

Fr att verfra Snabbstartsvrdet till apparaten och terg till Programmeny, tryck p NERT pil (Finish). Tryck sedan p VNSTER pil komma tillbaka till Mtlget. tre gnger fr att

1.1.3. Snabbstart Flde


Anvnd detta hjlpprogram fr att konfigurera enkla fldestillmpninger. [Visas p LUT430 (Pump och and Flde), och LUT440 (OCM) endast p konfigurerade modeller.] a) Tryck in HGER pil MENU. b) Tryck in HGER pil fr att aktivera PROGRAM-lge och ppna menyniv 1: MAIN tv gnger fr att navigera till menyposition 1.1.1. fr att ppna

c) Tryck p NERT pil tv gnger och sedan p HGER pil Snabbstartens hjlpprogram fr Flde. d) Vid varje steg, tryck p NERT pil fr att acceptera standardvrden och g direkt till nsta position, eller tryck p HGER pil fr att ppna funktionen Redigera: det aktuella valet r upplyst. e) Rulla till den nskade positionen och tryck p HGER pil fr att spara ndringen, och fr att fortstta. tryck sedan p NERT pil f)
QUICK START FLOW

TRANSDUCER XRS-5
BACK CANCEL NEXT EDIT

QUICK START FLOW


TRANSDUCER

NO TRANSDUCER XPS-10 XPS-15 XCT-8 XCT-12

XRS-5

Du kan nr som helst trycka p UPPT pil pil fr att annullera hjlpprogrammet. Start av Snabbstartens Fldeshjlpprogram

fr att g tillbaka, eller p VNSTER

Visar den typ av hjlpprogram som ska kras.

Alternativ

CANCEL, START

7ML19985XU81

SITRANS LUT400 (HART)

Sida SV-23

Transduktor

Specifierar vilken Siemens transduktor som r ansluten till apparaten.

Svenska

Alternativ
Temperaturklla

NO TRANSDUCER, XRS-5, XPS-10, XPS-15, XCT-8, XCT-12, XPS-30, XPS-40, XLT-30, XLT-60, STH Standard: NO TRANSDUCER

Avlsning av temperaturkllan anvnds fr att justera ljudets hastighet.

Alternativ

TRANSDUCER, FIXED TEMPERATURE, EXTERNAL TS-3, AVERAGE OF SENSORS Standard: TRANSDUCER

Faststlld Temperatur

Avlsning av temperaturkllan anvnds fr att justera ljudets hastighet.

Vrde

Omrde: -100,0 till +150,0 C Standard: +20,0 C

Denna parameter visas endast om FIXED TEMPERATURE valts som Temperaturklla. Primr Mtapparat

Definierar vilken primr mtapparat (PMD) som ska anvndas i tillmpningen.


EXPONENTIAL DEVICES (EXPONENTIELLA APPARATER), RECTANGULAR FLUME BS-3680 (REKTANGULR RNNA), ROUND NOSE HORIZONTAL CR. BS-3680 (HORISONTELL RUND SPETS). TRAPEZOIDAL FLUME BS3680 (TRAPETSFORMAD RNNA), U-FLUME BS-3680 (U-RNNA), FINITE CREST WEIR BS-3680 (NDLIG TOPPFRDMNING), THIN PLATE RECT. WEIR BS-3680 (REKT. TUNNPLTSFRDMNING), THIN PLATE V-NOTCH WEIR BS-3680 (TUNNPLTSFRDMNING V-SKRA BS-3680), RECT. WEIR CONTRACTED (REKT. FRDMNING SAMMANTRNGD), ROUND PIPE (RUNT RR), PALMER BOWLUS FLUME (PALMER-BOWLUS RNNA), H-FLUME (H-RNNA), OTHER (ANNAT)* Standard: EXPONENTIAL DEVICES

Alternativ

* Alternativet stlls p OTHER (ANNAT) om hjlpprogrammet tidigare krts via HART programverktyg (s som SIMATIC PDM), och apparaten stllts in p OFF eller UNIVERSAL HEAD VS. FLOW (Universell topp kontra flde). Om detta r den ursprungliga konfigurationen, kan PMD endast stllas p ej berkning (OFF), eller p linearisation (UNIVERSAL HEAD VS. FLOW) via HART programvaruverktyg (SIMATIC PDM, AMS, FC375/475). Se fullstndiga Driftsinstruktioner. Metod fr Fldesberkning

Stller in metod fr fldesberkning.

Alternativ
Enheter

ABSOLUTE (ABSOLUT), RATIOMETRIC (RATIOMETRISK) Standard: ABSOLUTE

Sensormtenheter.

Alternativ

M, CM, MM, FT, IN Standard: M

Anmrkning: I detta exempel antas alla vrden vara i meter (m).


Sida SV-24 SITRANS LUT400 (HART) 7ML19985XU81

vre kalibreringspunkt

Avstnd frn Sensornollpunkt till vre kalibreringspunkt: vanligen process full niv..

Vrde

Omrde: 0,000 till 60,000 Standard: 0,000

Svenska

Undre kalibreringspunkt

Avstnd frn Sensornollpunkt till Undre kalibreringspunkt: vanligen process tom niv.

Vrde
Svarshastighet

Omrde: 0,000 till 60,000 Standard: 60,000

Stller in enhetens reaktionshastighet p mtvrdesndringar i mlomrdet.

Anmrkningar:

Svarshastigheten kan endast stllas in genom Snabbstartsguiden, och alla ndringar som grs i Fill Rate per Minute (2.3.1.) eller Empty Rate per Minute (2.3.2.) parametrarna efter fullbordandet av hjlpprogrammet kommer att trnga bort instllningen fr Svarshastighet. Svarshastighet visas alltid i m/minut.
SLOW (0,1 M/MIN) MEDIUM (1,0 M/MIN) FAST (10 M/MIN) Standard: SLOW (0,1 M/MIN)

Alternativ

Anvnd en instllning just ver den maximala fyllnings- eller tmningshastigheten (den som r strst av dem). Lngsammare instllningar ger hgre precision; snabbare instllningar ger strre nivvariationer. Berkningsfaktorer:

Anmrkningar:

Fljande fem parametrar kommer att visas i hjlpprogrammet baserat p den PMD som valts ovan. Dessa parametrar anvnds i fldesberkningsformeln (se fullstndiga Driftsinstruktioner fr mer detaljer). Fldesexponent (PMD = EXPONENTIAL DEVICES (Exponentiella apparater))

Exponenten fr fldesberkningsformeln.

Vrde

Omrde: -999,000 till 9999,000 Standard: 1,550

7ML19985XU81

SITRANS LUT400 (HART)

Sida SV-25

K-Faktor (PMD = EXPONENTIAL DEVICES (Exponentiella apparater))

Svenska

Konstant som anvnds i fldesberkningsformeln endast fr absolut berkning av en exponentiell apparat.

Vrde
V-Skra Vinkel

Omrde: -999,000 till 9999,000 Standard: 1,000

(PMD = THIN PLATE V-NOTCH WEIR (Tunnpltsfrdmning v-skra))

V-Skrevinkeln som anvnds i fldesberkningsformeln.

Vrde
Lutning

Omrde: 25,000 till 95,000 Standard: 25,000

(PMD = TRAPEZOIDAL FLUME (Trapetsformad rnna) eller ROUND PIPE (Runt rr))

Den fldeslutning som anvnds i fldesberkningsformeln.

Vrde

Omrde: -999,000 till 9999,000 Standard: 0,000

Ytjmnhetskoefficient (PMD = ROUND PIPE (Runt rr))

Fldets Ytjmnhetskoefficient i fldesberkningsformeln.

Vrde

Omrde: -999,000 till 9999,000 Standard: 0,000

Sida SV-26

SITRANS LUT400 (HART)

7ML19985XU81

PMD Dimensioner

Anmrkningar:

Fr varje PMD som utesluter Exponential Devices (Exponentiella Apparater), och Other (Annat), mste du fra in upp till fyra dimensioner. I hjlpprogrammet, kommer du att tillfrgas om varje ndvndig dimension fr vald PMD, och respektive PMD dimensions namn kommer att visas. Vald PMD Dimensionens namn i hjlpprogrammet (referens parametermeny)
APPROACH WIDTH B (Instllningsbredd B) (2.15.4.5. OCM Dimension 1) THROAT WIDTH B (Halsbredd B) (2.15.4.6. OCM Dimension 2) HUMP HEIGHT P (Puckelhjd P) (2.15.4.7. OCM Dimension 3) THROAT LENGTH L (Halslngd L) (2.15.4.8. OCM Dimension 4)

Svenska

Rectangular Flume BS-3680 (Rektangulr rnna)

Round Nose Horizontal Crest Weir BS-3680 (Horisontell RundspetsToppfrdmning)


CREST WIDTH B (Toppbredd B) (2.15.4.5. OCM Dimension 1) CREST HEIGHT P (Topphjd P) (2.15.4.6. OCM Dimension 2) CREST LENGTH L (Topplngd L) (2.15.4.7. OCM Dimension 3)

Trapezoidal Flume (Trapetsformad rnna) BS-3680


APPROACH WIDTH B (Instllningsbredd B) (2.15.4.5. OCM Dimension 1) THROAT WIDTH B (Halsbredd B) (2.15.4.6. OCM Dimension 2) HUMP HEIGHT P (Puckelhjd P) (2.15.4.7. OCM Dimension 3) THROAT LENGTH L (Halslngd L) (2.15.4.8. OCM Dimension 4)

U-Flume BS-3680 (U-rnna)


APPROACH DIAMETER DA (Instllningsdiameter) (2.15.4.5. OCM Dimension 1) THROAT DIAMETER D (Halsdiameter) (2.15.4.6. OCM Dimension 2) HUMP HEIGHT P (Puckelhjd) (2.15.4.7. OCM Dimension 3) THROAT LENGTH L (Halslngd) (2.15.4.8. OCM Dimension 4)

Finite Crest Weir BS-3680 (Begrnsad Toppfrdmning)


CREST WIDTH B (Toppbredd) (2.15.4.5. OCM Dimension 1) CREST HEIGHT P (Topphjd) (2.15.4.6. OCM Dimension 2) CREST LENGTH L (Topplngd) (2.15.4.7. OCM Dimension 3)

Thin Plate Rectangular Weir BS-3680 (Rektangulr Tunnpltsfrdmning)


APPROACH WIDTH B (Instllningsbredd) (2.15.4.5. OCM Dimension 1) CREST WIDTH B (Toppbredd) (2.15.4.6. OCM Dimension 2) CREST HEIGHT P (Topphjd) (2.15.4.7. OCM Dimension 3)

Rectangular Weir Contracted (Rektangulr frdmning sammandragen)


CREST WIDTH B (Toppbredd) (2.15.4.5. OCM Dimension 1) (forts. p nsta sida)

7ML19985XU81

SITRANS LUT400 (HART)

Sida SV-27

PMD vald (forts.) Round Pipe (Runt rr)

Dimensionsnamn hjlpprogam (referens menyparameter)


PIPE INSIDE DIAMETER D (Diameter Rrets Insida) (2.15.4.5. OCM Dimension 1)

Svenska

Palmer Bowlus Flume (Palmer Bowlus rnna)


MAXIMUM FLUME WIDTH HMAX (Maximal rnnbredd) (2.15.4.5. OCM Dimension 1)

H-Flume (H-rnna)
MAXIMUM LISTED HEAD HMAX (Maximal Registrerad Niv) (2.15.4.5. OCM Dimension 1)

Maximal Niv

Maximalt nivvrde frbundet med PMD.

Vrde
Noll Nivmotvikt

Omrde: 0,000 till 60,000 Standard: 60,000

Skillnaden (positiv eller negativ) mellan Undre kalibreringspunkt och noll niv (niv vid noll flde).

Vrde

Omrde: -60,000 till 60,000 Standard: 0,000

Fldeshastighet enheter

Volymenheter som anvnds fr att visa totalt flde.

Alternativ

L/S, L/MIN, CUFT/S, CUFT/D, GAL/MIN, GAL/D, IMPGAL/MIN, IMPGAL/D, CUM/H, CUM/D, USER DEFINED * Standard: L/S

* Om USER DEFINED alternativ valts, mste vrdet regleras efter det att hjlpprogrammet fullfljts. Se parameter 2.6.6.User Defined Unit i de fullstndiga Driftsinstruktionerna. Maximalt Flde vid 20 mA

Maximal fldeshastighet.

Vrde

Omrde: -999 till 9999999 Standard: 100

Fldeshastighet Decimal

Maximalt antal visade decimalenheter.

Alternativ

NO DIGITS (INGEN SIFFRA), 1 DIGIT (1 SIFFRA), 2 DIGITS (2 SIFFROR), 3 DIGITS (3 SIFFROR) Standard: NO DIGITS

Avstngning Flde

Eliminerar totalisatoraktivitet fr niver vid eller under avstngningsvrdet.

Vrde

Omrde: 0,000 till 60,000 Standard: 0,000

Sida SV-28

SITRANS LUT400 (HART)

7ML19985XU81

Slut p Snabbstartens Fldeshjlpprogram

Fr att Snabbstarten ska lyckas, mste alla ndringar tillmpas.


BACK, CANCEL, FINISH (Visning tergr till 1.1.1 Snabbstartsguidens meny nr

Svenska

Alternativ Snabbstarten r genomfrd med framgng eller annullerad.) Om CANCEL


valts, frs inga ndringar in i apparaten.)

Fr att verfra Snabbstartsvrdena till apparaten och terg till Programmeny, tryck p NERT pil (Finish). Tryck sedan p VNSTER pil tre gnger fr att komma tillbaka till Mtlget.

Anmrkning: Vi rekommernderar starkt att en 2.15.2.Auto Zero Head utfrs efter

det att hjlprogrammet genomfrts fr att frskra dig om hgsta precision. Se de fullstndiga Driftsinstruktionerna fr mer detaljer.

1.2. Pump Styrning


Anvnd detta hjlpprogram fr att konfigurera pumparna om sdana ska anvndas i din tillmpning. Glm inte att frst genomfra tillmplig Snabbstart. a) Tryck in HGER pil MENU. b) Tryck in HGER pil c) Tryck p NERT pil fr att aktivera PROGRAM-lge och ppna menyniv 1: MAIN fr att navigera till menyposition 1.1. och HGER pil fr att ppna Pumpstyrningen. fr att acceptera standardvrden och g

d) Vid varje steg, tryck p NERT pil direkt till nsta position, eller HGER pil

fr att ppna funktionen Redigera: det aktuella valet r upplyst. fr att spara ndringen,

e) Rulla till den nskade positionen och tryck p HGER pil och tryck sedan p NERT pil f) fr att fortstta.

Du kan nr som helst trycka p UPPT pil pil fr att annullera hjlpprogrammet. Start av hjlpprogrammet - Pumpstyrning

fr att g tillbaka, eller VNSTER

Visar den typ av hjlpprogam som ska kras.

Alternativ
Antal Pumpar

CANCEL, START

Vlj det antal pumpar som ska anvndas vid pumpstyrningen.

Alternativ

NONE (INGEN), 2 Standard: NONE

Om den r instlld p NONE, r pumpstyrningen avaktiverad.

7ML19985XU81

SITRANS LUT400 (HART)

Sida SV-29

Rel Pump 1

Vljer den rel som r tilldelad Pump 1.

Svenska

Alternativ
Rel Pump 2

RELAY 2 (REL 2), RELAY 3 (REL 3) Standard: RELAY 2

Endast bild. Stller automatiskt in den rel som tilldelats Pump 2 enligt den rel som valts fr Pump 1 i fregende steg.
Om Rel Pump 1 = REL 2, d r Rel Pump 2 = REL 3 Alternativ (endast bild) Om Rel Pump 1 = REL 3, d r Rel Pump 2 = REL 2

Pumpstyrlge

Stller in den styrningsalgorism som anvnds fr att utlsa reln.

LUT420 Nivstyrenhet:
ALTERNATE DUTY ASSIST, ALTERNATE DUTY BACKUP

LUT430 Pump- och Fldesstyrenhet: Mjliga alternativ per modell


ALTERNATE DUTY ASSIST, ALTERNATE DUTY BACKUP , SERVICE RATIO DUTY ASSIST, SERVICE RATIO DUTY BACKUP , FIXED DUTY ASSIST, FIXED DUTY BACKUP

LUT440 Hg Precision OCM:


ALTERNATE DUTY ASSIST, ALTERNATE DUTY BACKUP , SERVICE RATIO DUTY ASSIST, SERVICE RATIO DUTY BACKUP , FIXED DUTY ASSIST, FIXED DUTY BACKUP Standard (alla modeller): ALTERNATE DUTY ASSIST

Service Ratio Pump 1

Vljer pumpanvndning enligt KRtidsratio snarare n senast anvnd.

Vrde

Omrde: 0 till 255 Standard: 1

Denna parameter visas endast om en Service Ratio algorism har valts fr Pumpstyrlge. Service Ratio Pump 2

Vljer pumpanvndning enligt KRtidsratio snarare n senast anvnd .

Vrde

Omrde: 0 till 255 Standard: 1

Denna parameter visas endast om en Service Ratio algorism har valts fr Pumpstyrlge.

Sida SV-30

SITRANS LUT400 (HART)

7ML19985XU81

KrTid Rel 2

Fr in den tid som pump Rel 2 har krt, definierad i timmar.

Vrde

Omrde: 0 till 999999 Standard: 0

Svenska

Anvnd standardvrdet fr nya pumpar, eller fr in vrdet fr existerande pumpar med sammanlagd krtid. (Denna parameter visas endast om en Service Ratio algorism har valts fr Pumpstyrlge.) Krtid Rel 3

Fr in den tid som pump Rel 3 har krt, definierad i timmar.

Vrde

Omrde: 0 till 999999 Standard: 0

Anvnd standardvrdet fr nya pumpar, eller fr in vrdet fr existerande pumpar med sammanlagd krtid. (Denna parameter visas endast om en Service Ratio algorism har valts fr Pumpstyrlge.) ON Vrde Pump 1

Den niv vid vilken Pump 1 stts p ON, definierad i 2.1.1.Units.

Vrde

Omrde: 0 till 99999,000 Standard: 0.000

ON Vrde Pump 2

Den niv vid vilken Pump 2 stts p ON, definierad i 2.1.1.Units.

Vrde

Omrde: 0,000 till 99999,000 Standard: 0,000

OFF Vrde Pump 1

Den niv dr Pump 1 stnger av OFF, definierad i 2.1.1.Units.

Vrde

Omrde: 0,000 till 99999,000 Standard: 0,000

OFF Instllningspunkt Pump 2

Den niv dr Pump 2 stnger av OFF, definierad i 2.1.1.Units.

Vrde

Omrde: 0,000 till 99999,000 Standard: 0,000

Slut p hjlpprogrammet - Pumpstyrning

Fr att hjlpprogrammet ska lyckas, mste alla ndringar tillmpas.


BACK, CANCEL, FINISH (Displayen tergr till Pumpstyrningsmenyn nr

Alternativ hjlpprogrammet gtts igenom med framgng eller annullerats. Om CANCEL


valts, frs inga ndringar in i apparaten.)

Fr att verfra vrdena till apparaten och terg till Programmenyn, tryck p NERT pil till Mtlget. (Finish). Tryck sedan p VNSTER pil tv gnger fr att g tillbaka

7ML19985XU81

SITRANS LUT400 (HART)

Sida SV-31

SITRANS LUT400 Kommunikationer: HART


SITRANS LUT400 kan konfigureras via HART-ntverket med hjlp av HART Communicator 375/475 frn Emerson (se de fullstndiga Driftsinstruktionerna fr mer detaljer), eller programvaror. De rekommenderade programvarorna r SIMATIC Process Device Manager (PDM) frn Siemens. Detaljerna skildras i grova drag i en tillmpningsguide Working with HART (Arbeta med HART), som kan laddas ner frn vr webbplats produktsida. G till: www.siemens.com/sitransLUT400 under Support och klicka p Application Guides (Tillmpningsguider).

Svenska

Fjrrdrift
SITRANS LUT400 stder flera programvaruverktyg fr drift via fjrrkommunikationer: PC som kr SIMATIC PDM PC som kr Emerson AMS Device Manager PC som kr en webbskare PC som kr Field Device Tool (FDT) Field Communicator 375/475 (FC375/FC475). Nedan hittar du ndvndig grundlggande information fr att anvnda din SITRANS LUT400. Fler detaljer fr varje verktyg finns tillgngliga i Kommunikationer fr SITRANS LUT400 (HART) Manualen.

SIMATIC PDM
SIMATIC PDM r ett programvarupaket som anvnds fr att kra igng och upprtthlla funktionen hos SITRANS LUT400 och andra processapparater. Du finner mer information p www.siemens.com/simatic-pdm. Kontrollera support-sidan p vr webbplats fr att skerstlla att du har den senaste versionen av SIMATIC PDM, senaste Service Pack (SP) och senaste hot fix (HF). G till: http://support.automation.siemens.com/WW/ llisapi.dll?func=cslib.csinfo&lang=en&objiD=10806857&subtype=133100

Uppdatera Electronic Device Description (EDD) (Beskrivning Elektroniska Instrument)


Du hittar EDD i Instrumentkatalogen Device Catalog, under Sensors/Level/Echo/Siemens AG/ SITRANS LUT400. (EDD r skriven fr att frmja kompatibilitet.) Fr att hjlpa till att hitta korrekt EDD, ska major och minor siffror passa mellan EDD revisionen och Firmwarerevisionen i apparaten (t.ex. major och minor siffror i fetstil: 1.00.00-02).
EDD-revisionen mste stmma med apparatens Firmwarerevisioner. Fr att kontrollera den i PDM, g till SITRANS LUT400 HART > Identification (Identifiering) > Device (Apparat).

passande Firmware och EDDrevisioner

Installering av en ny version av SIMATIC PDM fordrar den senaste Service Pack (SP) och den senaste hot fix (HF).

Fr att installera en ny EDD


G till www.siemens.com/sitransLUT400 > Support > Software Downloads fr att ladda ner den senaste EDD frn produktsidan p vr webbplats. Spara filerna p din dator, och extrahera den zippade filen till en ltttkomlig plats. Starta SIMATIC PDM Manage Device Catalog (Instrumentkatalog), blddra till den uppackade EDD-filen och vlj den.

Sida SV-32

SITRANS LUT400 (HART)

7ML19985XU81

Konfigurering av en ny apparat
1) 2) Kontrollera att du har den senaste EDD:n, och uppdatera den vid behov (seFr att installera en ny EDD, ovan). Konfigurera SITRANS LUT400 med hjlp av Snabbstartsguiden fr din tillmpning.

Svenska

Snabbstartsguiden via SIMATIC PDM


Anmrkningar:
Snabbstartsguidens instllningar r beroende av varandra och ndringar gller frst efter att du klickat p Apply and Transfer (Tillmpa och verfr) i slutet av det sista steget. Klicka p BACK (TILLBAKA) fr att g tillbaka och gra om en instllning eller CANCEL fr att lmna Snabbstarten. ppna Device (Apparat) menyn i SIMATIC PDM och vlj Snabbstart fr din tillmpning (Niv, Volym, Volym-Linearisering, Flde).

Snabbstart

Funktion via AMS Device Manager


AMS r ett programvarupaket som anvnds fr att kra igng och upprtthlla funktionen hos SITRANS LUT400 och andra processapparater. Mer information ges p: http://www.emersonprocess.com/AMS/.

Electronic Device Description (EDD) (Beskrivning Elektroniska Instrument)


SITRANS LUT400 krver EDD fr AMS Device Manager version 10.5.

Konfiguration av en ny apparat via AMS Device Manager


1) Du hittar EDD i Instrumentkatalogen Device Catalog, under Sensors/Level/Echo/Siemens AG/SITRANS LUT400. Kontrollera produktsidan p vr webbplats www.siemens.com/sitransLUT400, under Support->Software Downloads, fr att frskra dig om att du har den senaste EDD versionen fr AMS Device Manager. Spara filerna p din dator, och extrahera den zippade filen till en ltttkomlig plats. Hmta in AMS Device Manager Add Device Type (AMS Apparatanvisning Lgg till apparattyp), blddra till den uppackade EDD-filen och vlj den. (En Applikationsvgledning fr installation av HART-apparater med AMS Device Manager kan laddas ner frn produktsidan p vr webbplats under Support.) I Device Connection View (Apparatanslutning Bild), hgerklicka p apparatikonen och vlj Scan Device fr att ladda upp parametrar frn apparaten. Dubbelklicka p apparatikonen fr att ppna startskrmbilden. Startskrmbilden visar enhetens identifikationsdetaljer, och ett navigationsfnster till vnster p skrmen. Konfigurera SITRANS LUT400 med hjlp av lmplig Snabbstartsguide.

2) 3)

4) 5) 6)

7ML19985XU81

SITRANS LUT400 (HART)

Sida SV-33

Funktion via FC375/475 (HART)


Svenska
FC375/FC475 HART Communicator r en kommunikationsapparat du kan hlla i handen, som r ltt att avnda, och ger universellt std fr andra HART-apparater, s som SITRANS LUT400. FC375/475s menystruktur r mycket lik den i AMS Device Manager.

Electronic Device Description (EDD) (Beskrivning Elektroniska Instrument)


SITRANS LUT400 fordrar EDD fr FC375/FC475.

Konfigurering av en ny apparat FC375/FC475


1) Se produktsidan p vr webbplats p www.siemens.com/sitransLUT400, under Support>Software Downloads, fr att vara sker p att du har den senaste versionen av EDD fr FC375/FC475. Spara filerna p din dator, och extrahera den zippade filen till en ltttkomlig plats. Fr att installera EDD, kr Emersons Field Communicator Easy Upgrade Utility. Konfigurera SITRANS LUT400 med hjlp av Snabbstartsguide fr din tillmning.

2) 3) 4)

Funktion via FDT (Field Device Tool [Fltapparatverktyg])


FDT r en standard som anvnds i mnga programvarupaket avsedda fr att kra igng och upprtthlla fltapparatur som SITRANS LUT400. Tv kommersiellt tillgngliga FDT:n r PACTware och Fieldcare. FDT r mycket lik PDM. Fr att konfigurera en fltapparat via FDT behver du DTM (Device Type Manager) fr enheten. Siemens-instrument anvnder SITRANS DTM och ett EDDinstrument skrivet fr SITRANS DTM. 1) Installera SITRANS DTM i ert system. Du kan ladda ner det frn: http://www.siemens.com/sitransdtm. Klicka p Support g sedan till Software downloads. Installera SITRANS LUT400 (HART) EDD fr SITRANS DTM. Den kan laddas ner frn produktsidan p vr webbplats: G till www.siemens.com/sitransLUT400 under Support och klicka p Software Downloads.

2)

Konfigurering av en ny apparat via FDT


En Applikationsvgledning kan laddas ner frn produktsidan p vr webbplats under Support.

Service och Underhll


SITRANS LUT400 krver vare sig underhll eller rengring under normala driftfrhllanden.

Reparation och ansvarsbegrnsning


Fr detaljerad information, v.g. se omslagets tredje sida.

Sida SV-34

SITRANS LUT400 (HART)

7ML19985XU81

Kabeldragningar fr installationer i farligt omrde


Fljande kabeldragningsalternativ finns tillgngliga fr installationer i farligt omrde: Dammantndningsker kabling (Internationelltl/Europa) Icke antndbar kabling (Kanada) Kontrollera alltid instrumentets mrkskylt, och bekrfta angivna vrden.

Svenska

1. Dammantndningsker kabling (Internationellt/Europa)


SITRANS LUT420 7MLxxxx-xxxxx-xxxx Serial No.: GYZ / B1034567 0518 KCC-REM-S49 Power Rating: 100 230V 15% SITRANSLUT400 50/60 Hz, 36 VA (10 W) Contact Rating: 1A/5A @ 250V , Non-Inductive Operating Temperature: 20C to 50C Enclosure: IP65 / TYPE 4X / NEMA 4X II 3D Ex tc IIIC T70C Dc Ex tc IIIC T70C Dc SIRA 12ATEX9143X IECEx SIR 12.0057X Warning: Possible static hazard. Do not rub or clean on site with a dry cloth. Warning: Do not open when energized or when an explosive dust atmosphere is present. Siemens Milltronics Process Instruments, Peterborough Assembled in Canada with domestic and imported parts SITRANS LUT430 7MLxxxx-xxxxx-xxxx Serial No.: GYZ / B1034567 0518 KCC-REM-S49 Power Rating: 100 230V 15% SITRANSLUT400 50/60 Hz, 36 VA (10 W) Contact Rating: 1A/5A @ 250V , Non-Inductive Operating Temperature: 20C to 50C Enclosure: IP65 / TYPE 4X / NEMA 4X II 3D Ex tc IIIC T70C Dc Ex tc IIIC T70C Dc SIRA 12ATEX9143X IECEx SIR 12.0057X Warning: Possible static hazard. Do not rub or clean on site with a dry cloth. Warning: Do not open when energized or when an explosive dust atmosphere is present. Siemens Milltronics Process Instruments, Peterborough Assembled in Canada with domestic and imported parts SITRANS LUT440 7MLxxxx-xxxxx-xxxx Serial No.: GYZ / B1034567 0518 KCC-REM-S49 Power Rating: 100 230V 15% SITRANSLUT400 50/60 Hz, 36 VA (10 W) Contact Rating: 1A/5A @ 250V , Non-Inductive Operating Temperature: 20C to 50C Enclosure: IP65 / TYPE 4X / NEMA 4X II 3D Ex tc IIIC T70C Dc Ex tc IIIC T70C Dc SIRA 12ATEX9143X IECEx SIR 12.0057X Warning: Possible static hazard. Do not rub or clean on site with a dry cloth. Warning: Do not open when energized or when an explosive dust atmosphere is present. Siemens Milltronics Process Instruments, Peterborough Assembled in Canada with domestic and imported parts

SITRANS LUT420 7MLxxxx-xxxxx-xxxx Serial No.: GYZ / B1034567 0518 KCC-REM-S49 Power Rating: 10 32V , 10W SITRANSLUT400 Contact Rating: 1A/5A @ 250V , Non-Inductive Operating Temperature: 20C to 50C Enclosure: IP65 / TYPE 4X / NEMA 4X

SITRANS LUT430 7MLxxxx-xxxxx-xxxx Serial No.: GYZ / B1034567 0518 KCC-REM-S49 Power Rating: 10 32V , 10W SITRANSLUT400 Contact Rating: 1A/5A @ 250V , Non-Inductive Operating Temperature: 20C to 50C Enclosure: IP65 / TYPE 4X / NEMA 4X

SITRANS LUT440 7MLxxxx-xxxxx-xxxx Serial No.: GYZ / B1034567 0518 KCC-REM-S49 Power Rating: 10 32V , 10W SITRANSLUT400 Contact Rating: 1A/5A @ 250V , Non-Inductive Operating Temperature: 20C to 50C Enclosure: IP65 / TYPE 4X / NEMA 4X

II 3D Ex tc IIIC T70C Dc Ex tc IIIC T70C Dc SIRA 12ATEX9143X IECEx SIR 12.0057X Warning: Possible static hazard. Do not rub or clean on site with a dry cloth. Warning: Do not open when energized or when an explosive dust atmosphere is present.

II 3D Ex tc IIIC T70C Dc Ex tc IIIC T70C Dc SIRA 12ATEX9143X IECEx SIR 12.0057X Warning: Possible static hazard. Do not rub or clean on site with a dry cloth. Warning: Do not open when energized or when an explosive dust atmosphere is present.

II 3D Ex tc IIIC T70C Dc Ex tc IIIC T70C Dc SIRA 12ATEX9143X IECEx SIR 12.0057X Warning: Possible static hazard. Do not rub or clean on site with a dry cloth. Warning: Do not open when energized or when an explosive dust atmosphere is present.

Siemens Milltronics Process Instruments, Peterborough Assembled in Canada with domestic and imported parts

Siemens Milltronics Process Instruments, Peterborough Assembled in Canada with domestic and imported parts

Siemens Milltronics Process Instruments, Peterborough Assembled in Canada with domestic and imported parts

ATEX-intyg listade p mrkskylten kan laddas ner frn produktsidan p vr webbplats p www.siemens.com/sitransLUT400. IECEx-intyget listat p mrkskylten kan ses p IECEx-webbplatsen. G till: http://iecex.iec.ch och klicka p Ex Equipment Certificates of Conformity (verensstmmelseintyg fr utrustning) och fr sedan in intygsnumret IECEx SIR 12.0057X Fr kabeldragningskrav, flj lokala bestmmelser. Se ven ATEX-intyg listat ovan.

7ML19985XU81

SITRANS LUT400 (HART)

Sida SV-35

2. Icke antndbar kabling (Kanada)


Svenska
SITRANS LUT420 7MLxxxx-xxxxx-xxxx Serial No.: GYZ / B1034567 Power Rating: 100 230V 15% 50/60 Hz, 36 VA (10 W) Contact Rating: 1A/5A @ 250V , Non-Inductive Operating Temperature: 20C to 50C Enclosure: IP65 / TYPE 4X / NEMA 4X SITRANS LUT430 7MLxxxx-xxxxx-xxxx Serial No.: GYZ / B1034567 Power Rating: 100 230V 15% 50/60 Hz, 36 VA (10 W) Contact Rating: 1A/5A @ 250V , Non-Inductive Operating Temperature: 20C to 50C Enclosure: IP65 / TYPE 4X / NEMA 4X SITRANS LUT440 7MLxxxx-xxxxx-xxxx Serial No.: GYZ / B1034567 Power Rating: 100 230V 15% 50/60 Hz, 36 VA (10 W) Contact Rating: 1A/5A @ 250V , Non-Inductive Operating Temperature: 20C to 50C Enclosure: IP65 / TYPE 4X / NEMA 4X

159134

Class I, Div.2, Gr. A, B, C & D T3C Class II, Div. 2, Gr. F & G Class III Per Drawing A5E03936871

159134

Class I, Div.2, Gr. A, B, C & D T3C Class II, Div. 2, Gr. F & G Class III Per Drawing A5E03936871

159134

Class I, Div.2, Gr. A, B, C & D T3C Class II, Div. 2, Gr. F & G Class III Per Drawing A5E03936871

Siemens Milltronics Process Instruments, Peterborough Assembled in Canada with domestic and imported parts

Siemens Milltronics Process Instruments, Peterborough Assembled in Canada with domestic and imported parts

Siemens Milltronics Process Instruments, Peterborough Assembled in Canada with domestic and imported parts

SITRANS LUT420 7MLxxxx-xxxxx-xxxx Serial No.: GYZ / B1034567 Power Rating: 10 32V , 10W Contact Rating: 1A/5A @ 250V , Non-Inductive Operating Temperature: 20C to 50C Enclosure: IP65 / TYPE 4X / NEMA 4X

SITRANS LUT430 7MLxxxx-xxxxx-xxxx Serial No.: GYZ / B1034567 Power Rating: 10 32V , 10W Contact Rating: 1A/5A @ 250V , Non-Inductive Operating Temperature: 20C to 50C Enclosure: IP65 / TYPE 4X / NEMA 4X

SITRANS LUT440 7MLxxxx-xxxxx-xxxx Serial No.: GYZ / B1034567 Power Rating: 10 32V , 10W Contact Rating: 1A/5A @ 250V , Non-Inductive Operating Temperature: 20C to 50C Enclosure: IP65 / TYPE 4X / NEMA 4X

159134

Class I, Div.2, Gr. A, B, C & D T3C Class II, Div. 2, Gr. F & G Class III Per DWG. A5E03936871

159134

Class I, Div.2, Gr. A, B, C & D T3C Class II, Div. 2, Gr. F & G Class III Per DWG. A5E03936871

159134

Class I, Div.2, Gr. A, B, C & D T3C Class II, Div. 2, Gr. F & G Class III Per DWG. A5E03936871

Siemens Milltronics Process Instruments, Peterborough Assembled in Canada with domestic and imported parts

Siemens Milltronics Process Instruments, Peterborough Assembled in Canada with domestic and imported parts

Siemens Milltronics Process Instruments, Peterborough Assembled in Canada with domestic and imported parts

CSA Klass I, Div 2 anslutning ritning nummer A5E03936871 kan laddas ner frn produktsidan p vr webbplats p www.siemens.com/sitransLUT400.

Instruktioner gllande installation i riskzoner (Referens Europeiska ATEX-direktivet 94/9/EG, Bilaga II, 1/0/6)
Fljande instruktioner r tillmpliga p utrustning som tcks av intyg nummer SIRA 12ATEX9143X: 1) 2) 3) 4) Fr anvndning och montering, se huvudinstruktionerna. Utrustningen r certifierad fr anvndning i Kategori 3D. Utrustningen kan anvndas i riskzon 22. Utrustningen har en maximal yttemperatur p T70 C (i en omgivning med +50C). Se tillmplig praxis fr val av utrustning med hnsyn till specifika dammantndningstemperaturer. 5) Utrustningen r certifierad fr anvndning inom ett temperaturomrde frn -20 till +50C. 6) Alla kabel- eller ledarinfringar mste uppfylla kraven i det Europeiska direktivet 94/9/EG fr Grupp II, Kategori 3D och upprtthlla den generella IP-klassningen fr kapslingen. 7) Kabel skall vljas i enlighet med tillmplig praxis och s att dess isolering kan motst kapslingens maximala yttemperatur (T70 C). 8) Utrustningen har inte bedmts som en skerhetsrelaterad utrustning (enligt meningen i direktiv 94/9/EG Bilaga II, klausul 1.5). 9) Installation och inspektion av denna utrustning skall utfras av lmpligt utbildad personal i enlighet med tillmpliga delar av arbetspraxis (EN 60079-14 och EN 60079-17 i Europa). 10) Reparation av denna utrustning skall utfras av lmpligt utbildad personal i enlighet med tillmpliga delar av arbetspraxis (t.ex. EN 60079-19 i Europa).

Sida SV-36

SITRANS LUT400 (HART)

7ML19985XU81

11) Komponenter som skall sttas in i eller anvndas som reservdelar till utrustningen skall monteras av lmpligt utbildad personal i enlighet med tillverkarens dokumentation. 12) Manuell frbikoppling kan utfras med hjlp av strmbrytare ngonstans i byggnadens elinstallation.

Svenska

SPECIELLA VILLKOR FR SKER ANVNDNING


X-suffixet i intygsnumret betecknar fljande speciella villkor fr sker anvndning: Alla kabel- eller ledningsinfrningar ska motsvara kraven i det Europeiska Direktivet 94/9/EG fr Grupp II, Kategori 3D och upprtthlla kapslingens generella IP klassning. Kabel ska vljas enligt tillmplig Praxis och dess isolering ska kunna motst kapslingens maximala yttertemperatur (T70 C). Kapslingen r icke-ledande, och kan under extrema frhllanden alstra en antndningsbar niv elektrostatisk laddning. Anvndaren ska frskra sig om att utrustningen inte r installerad p en plats dr den kan utstts fr extrema villkor (som t.ex. hgtrycksnga) som kan frorsaka bildning av elektrostatisk laddning p ickeledande ytor. Dessutom fr rengring endast ske med fuktig duk. Utrustningen ska antingen installeras i ett omrde med liten risk fr mekanisk skada eller ska den, om s inte r mjligt, frses med ytterligare en fr tillmpningen gnad skyddsanordning. Hela LUT400 seriens styrapparater (LUT420, LUT430, LUT440) som anvnds med distant panelmonterad lucka ska installeras i kapsling av passande klass som motsvarar kraven i det Europeiska Direktivet 94/9/EG fr Grupp II, Kategori 3D och vara klassad IP65 eller hgre.

7ML19985XU81

SITRANS LUT400 (HART)

Sida SV-37

Anteckningar
Svenska
Sida SV-38

SITRANS LUT400 (HART)

7ML19985XU81

Unit Repair and Excluded Liability

All changes and repairs must be done by qualified personnel, and applicable safety regulations must be followed. Please note the following: The user is responsible for all changes and repairs made to the device. All new components must be provided by Siemens. Restrict repair to faulty components only. Do not re-use faulty components

Reparation af enheden og ansvarsbegrnsning:


Alle ndringer og reparationer skal udfres af kvalificeret personale, og de gldende sikkerhedsbestemmelser skal overholdes. Bemrk venligst flgende: Brugeren er ansvarlig for alle de p apparatet udfrte ndringer og reparationer. Alle nye komponenter skal vre leveret af Siemens. Reparr kun defekte komponenter. Defekte komponenter m ikke genbruges

Gertereparatur und Haftungsausschluss:


Alle nderungen und Reparaturen mssen von qualifiziertem Personal unter Beachtung der jeweiligen Sicherheitsbestimmungen vorgenommen werden. Bitte beachten Sie: Der Benutzer ist fr alle nderungen und Reparaturen am Gert verantwortlich. Alle neuen Bestandteile sind von Siemens bereit zu stellen. Reparieren Sie lediglich defekte Bestandteile. Defekte Bestandteile drfen nicht wiederverwendet werden.

:
, . : . Siemens. . .

Reparacin del dispositivo y lmite de responsabilidad:


Las modificaciones y reparaciones debern ser efectuadas por personal calificado de acuerdo con las normas de seguridad aplicables. Notas importantes: El usuario es el nico responsable de las modificaciones y reparaciones del dispositivo. Recomendamos utilizar slo recambios originales Siemens. Reparar slo los componentes defectuosos. Los componentes defectuosos no se deben reutilizar.

Rparation de lunit et limite de responsabilit :


Les modifications et rparations doivent tre effectues par un personnel qualifi en accord avec les consignes de scurit applicables. Remarques importantes : Lutilisateur est seul responsable des modifications et rparations effectues sur lunit. Utiliser seulement des composants fournis par Siemens. Rparer uniquement les composants dfectueux. Les composants dfectueux ne doivent pas tre rutiliss.

Riparazioni dellapparecchiatura e limiti di responsabilit:


Le modifiche e le riparazioni devono essere effettuate solo da personale qualificato, rispettando le normative sulla sicurezza. Note importanti: Lutente responsabile delle eventuali modifiche e riparazioni effettuate sullapparecchiatura. Utilizzare solo pezzi di ricambio originali forniti da Siemens. Riparare solo i componenti difettosi. E importante non riutilizzare i componenti difettosi.

Reparatie van apparatuur en uitsluiting van aansprakelijkheid:


Alle modificaties en reparaties moeten worden uitgevoerd door gekwalificeerd personeel en de geldende veiligheidsvoorschriften moeten worden aangehouden. Let op: De gebruiker is verantwoordelijk voor alle modificaties en reparaties die worden uitgevoerd aan het apparaat. Alle nieuwe onderdelen moeten zijn geleverd door Siemens. Beperk de reparatie uitsluitend tot de defecte componenten. Defecte componenten niet opnieuw gebruiken.

Reparao da Unidade e Responsabilidade Excluda


Todas as alteraes e reparaes devem ser realizadas por pessoal qualificado e devem ser seguidas as regras de segurana aplicveis. Por favor, note o seguinte: O utilizador responsvel por todas as alteraes e reparaes efectuadas no dispositivo. Todos os novos componentes devem ser fornecidos pela Siemens. Reparao restrita apenas a componentes danificados. No reutilize componentes danificados.

Yksikn korjaaminen ja vastuuvapaus:


Muutos- ja korjaustyt saa suorittaa ainoastaan ptev henkilkunta, ja voimassa olevia turvallisuusmryksi on noudatettava. Pyydmme ottamaan huomioon seuraavat seikat: Kyttj on vastuussa kaikista laitteeseen tehdyist muutoksista ja korjauksista. Kaikki uudet osat on hankittava Siemens:lt. Korjaukset on kohdistettava ainoastaan viallisiin osiin. Viallisia osia ei saa kytt uudelleen.

Reparation och ansvarsfrihet:


Alla ndringar och reparationer mste utfras av kompetent personal och under iakttagande av gllande skerhetsbestmmelser. Observera att: Anvndaren ansvarar fr alla ndringar och reparationer som grs p enheten. Alla nya delar mste komma frn Siemens. Reparara endast med fel behftade delar. Delar behftade med fel fr ej teranvndas.

www.siemens.com/processautomation

For more information


www.siemens.com/level www.siemens.com/weighing

Siemens AG Industry Sector 1954 Technology Drive P.O. Box 4225 Peterborough, ON Canada K9J 7B1 email: techpubs.smpi@siemens.com

Subject to change without prior notice 7ML19985XU81 Rev. 1.0


Siemens AG 2012

*7ml19985XU81*
Printed in Canada

www.siemens.com/processautomation

You might also like