High Performance Ammonia Ion Selective Electrode

ROSS and the COIL trade dress are trademarks of Thermo Fisher Scientific Inc. AQUAfast, Cahn, ionplus, KNIpHE, No Cal, ORION, perpHect, PerpHecT, PerpHecTion, pHISA, pHuture, Pure Water, Sage, Sensing the Future, SensorLink, ROSS, ROSS Ultra, Sure-Flow, Titrator PLUS and TURBO2 are registered trademarks of Thermo Fisher. 1-888-pHAX-ION, A+, All in One, Aplus, AQUAsnap, AssuredAccuracy, AUTO-BAR, AUTO-CAL, AUTO DISPENSER, Auto-ID, AUTO-LOG, AUTO-READ, AUTO-STIR, Auto-Test, BOD AutoEZ, Cable-Free, CERTI-CAL, CISA, DataCOLLECT, DataPLUS, digital LogR, DirectCal, DuraProbe, Environmental Product Authority, Extra Easy/Extra Value, FAST QC, GAP, GLPcal, GLPcheck, GLPdoc, ISEasy, KAP, LabConnect, LogR, Low Maintenance Triode, Minimum Stir Requirement, MSR, NISS, One-Touch, One-Touch Calibration, One-Touch Measurement, Optimum Results, Orion Star, Pentrode, pHuture MMS, pHuture Pentrode, pHuture Quatrode, pHuture Triode, Quatrode, QuiKcheK, rf link, ROSS Resolution, SAOB, SMART AVERAGING, Smart CheK, SMART STABILITY, Stacked, Star Navigator 21, Stat Face, The Enhanced Lab, ThermaSense, Triode, TRIUMpH, Unbreakable pH, Universal Access are trademarks of Thermo Fisher. Guaranteed Success and The Technical Edge are service marks of Thermo Fisher. PerpHecT meters are protected by U.S. patent 6,168,707. PerpHecT ROSS are protected by U.S. patent 6,168,707. ORION Series A meters and 900A printer are protected by U.S. patents 5,198,093, D334,208 and D346,753. ionplus electrodes and Optimum Results solutions are protected by US Patent 5,830,338. ROSS Ultra electrodes are protected by US patents 6,793,787. Orion ORP Standard is protected by US Patent 6,350,367. Orion NoCal electrodes with stabilized potential patent pending. 2007 Thermo Fisher Scientific Inc. All rights reserved. All trademarks are the property of Thermo Fisher Scientific Inc. and its subsidiaries. The specifications, descriptions, drawings, ordering information and part numbers within this document are subject to change without notice. This publication supersedes all previous publications on this subject.

Table of Contents
General Information .. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 Required Equipment. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2

Electrode Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
Electrode Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 Electrode Preparation with Preassembled Outer Body and Membrane Cap Assembly . . . . . . . . . . . . . . . . 4 Electrode Preparation with Loose Membrane . . . . . . . . . . . . . . 6 Checking Electrode Operation (Slope) . . . . . . . . . . . . . . . . . . . . 9 Units of Measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10 Sample Requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10 Measuring Hints . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11 Sample Storage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12 Electrode Storage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12

Analytical Techniques .. . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Typical Calibration Curve. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Direct Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Low-Level Measurements By Direct Calibration . . . . . . . . . . . Measuring Ammonia in Solutions that Wet the Membrane . . . Measuring Organic Nitrogen . . . . . . . . . . . . . . . . . . . . . . . . . . Application Notes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Known Addition . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15 16 20 23 24 24 25

Electrode Characteristics .. . . . . . . . . . . . . . . . . . . . . . . . 31
Electrode Response . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Temperature Effects . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Interferences . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . pH Effects. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Complexation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Effect of Dissolved Species . . . . . . . . . . . . . . . . . . . . . . . . . . . Membrane Life . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Theory of Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31 32 32 33 33 33 33 34

Troubleshooting. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
Troubleshooting Checklist. . . . . . . . . . . . . . . . . . . . . . . . . . . . . Troubleshooting Guide . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Assistance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Warranty . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37 40 42 42

Ordering Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43 Specications .. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44

General Information
Introduction
The high performance ammonia ion selective electrode (ISE) allows fast, simple, economical, and accurate measurements of dissolved ammonia in aqueous solutions. This gas sensing electrode can also be used to measure the ammonium ion after conversion to ammonia or to measure organic nitrogen after Kjeldahl digestion of the sample. Sample color and turbidity do not affect the measurement and samples do not need to be distilled. Almost all anions, cations and dissolved species, other than volatile amines, do not interfere. All apparatus and solutions required for measurement, electrode characteristics and electrode theory are discussed in this user guide. General analytical procedures, low-level procedures and a method for measuring ammonia in solutions that wet the membrane are given. The High Performance Ammonia Electrode Includes: One preassembled outer body and membrane cap- includes outer body, membrane cap and membrane Twenty loose membranes Tweezers for handling loose membranes Electrode lling solution Dispensing cap for electrode lling solution User guide

There are two measurement techniques that are recommended for direct measurements of ammonia samples. The primary direct measurement technique is for ammonia samples with concentrations of 1 ppm ammonia as nitrogen and higher. The low-level measurement technique is for ammonia samples with concentrations of 1 ppm ammonia as nitrogen and lower. Refer to the Analytical Techniques section for details. Technical Support Chemists can be consulted for assistance and troubleshooting advice. Within the United States call 1.800.225.1480 and outside the United States call 978.232.6000 or fax 978.232.6031. In Europe, the Middle East and Africa, contact your local authorized dealer. For the most current contact information, visit www.thermo.com/water.

High Performance Ammonia Electrode User Guide

Required Equipment
1. Thermo Scientic Orion ISE meter, such as the 4-Star pH/ ISE meter or 5-Star pH/ISE/DO/conductivity meter. Ammonia electrodes can be used on any ISE meter with a BNC or U.S. standard connection. The electrodes can also be used on meters with a variety of inputs when an adapter cable is used. Visit www.thermo.com/water for details.

2. Thermo Scientic Orion high performance ammonia ion selective electrode, Cat. No. 9512HPBNWP or 9512HP01. 3. Magnetic stirrer or Orion stirrer probe, Cat. No. 096019. The Orion stirrer probe can be used with 3-Star, 4-Star and 5-Star benchtop meters. 4. Volumetric asks, graduated cylinders and beakers. 5. Distilled or deionized water. All water must be ammoniafree. Pass distilled or deionized water through an ionexchange column containing a strong acidic cation exchange resin, such as Dowex 50W-X8. 6. Ammonia electrode lling solution, Cat. No. 951209. 7 . Ammonia calibration standards. Cat. No. 951006 951007 951207 Description 0.1 M ammonia chloride standard 1000 ppm ammonia as nitrogen standard 100 ppm ammonia as nitrogen standard

8. Ionic Strength Adjuster (ISA), which provides a constant background ionic strength and adjusts the solution pH. Cat. No. 951211 951011 Description pH-adjusting ISA, for samples that have a concentration of 1 ppm as nitrogen or higher Alkaline reagent (low-level ISA), for samples that have a concentration of 1 ppm as nitrogen or lower

9. Ammonia electrode storage solution, 100 mL of 1 ppm standard with 1 mL of alkaline reagent (Cat. No. 951011). Refer to the Electrode Storage section for detailed instructions on storage solution preparation.

High Performance Ammonia Electrode User Guide

Electrode Setup
Electrode Assembly
Note: A new electrode is shipped without a membrane. You must install a preassembled outer body and membrane cap assembly or a new membrane before using the electrode. Note: A new electrode is shipped dry. You must soak the inner body of the electrode in the electrode lling solution for at least two hours before using the electrode. For best results, soak the inner body overnight in electrode lling solution. Assemble the electrode according to the instructions listed in the Electrode Preparation with a Preassembled Outer Body and Membrane Cap Assembly or Electrode Preparation with Loose Membranes section. Avoid excessive handling of the membrane during assembly; this may affect the hydrophobic properties of the membrane and cause a shortened membrane life. Use the tweezers provided. A membrane will last from one week to several months, depending on usage.

High Performance Ammonia Electrode User Guide

Electrode Preparation with a Preassembled Outer Body and Membrane Cap Assembly
1. Hold the electrode vertically and unscrew the electrode cap from the electrode body. See Figure 1. 2. Carefully remove the inner body from the outer body assembly. See Figure 2. Dispose of any electrode lling solution that is in the outer body. Note: A new electrode is shipped dry. You must soak the inner body of the electrode in the electrode lling solution for at least two hours before using the electrode. For best results, soak the inner body overnight in electrode lling solution.

ll

ll

ll

Figure 1

Figure 2

Figure 3

High Performance Ammonia Electrode User Guide

3. Select a new preassembled outer body and membrane cap assembly and verify that the membrane is not wrinkled or torn. See Figure 3. 4. Fill the preassembled outer body and membrane cap to the ll line with electrode lling solution. See Figure 4. 5. Insert the inner body into the preassembled outer body and membrane cap. Ensure that the inner body is fully inserted in the top of the outer body and then screw on the electrode cap. See Figure 5. 6. Gently tap the side of the electrode to remove air bubbles. 7 . Soak the electrode in the ammonia electrode storage solution for at least 15 minutes before use. Prepare the storage solution by combining 100 mL of 1 ppm ammonia standard with 1 mL of alkaline reagent, Cat. No. 951011.

g lin n l lutio o s

ll

ll

Figure 4

Figure 5

High Performance Ammonia Electrode User Guide

Electrode Preparation with Loose Membrane


1. Hold the electrode vertically and unscrew the electrode cap from the electrode body. See Figure 6. 2. Carefully remove the inner body from the outer body assembly. See Figure 7. Dispose of any electrode lling solution that is in the outer body. Note: A new electrode is shipped dry. You must soak the inner body of the electrode in the electrode lling solution for at least two hours before using the electrode. For best results, soak the inner body overnight in electrode lling solution.

3. Unscrew the membrane cap from the outer body. See Figure 8. Remove the membrane from the membrane cap if a membrane was previously installed.

ll

ll

ll

Figure 6

Figure 7

Figure 8

High Performance Ammonia Electrode User Guide

4. Wear gloves and use tweezers to carefully grasp the corner of the white membrane from between wax paper separators. See Figure 9. Note: Do not touch the center of the membrane.

5. Hold the outer body oriented with the smaller diameter threads at the top. See Figure 10. 6. Align the serrated edge of the membrane against the threaded shoulder and hold the membrane with your thumb. See Figure 11.

Figure 9 7 .

Figure 10

Figure 11

With the other hand, gently place the membrane across the opening. See Figure 12.

8. Then place the membrane down to align the other edge with the opposite shoulder. See Figure 13. 9. While holding each edge on both sides, gently place each smooth side of the membrane out and down over the threads and ensure that the membrane surface is smooth and without wrinkles. See Figure 14. 10. Smooth any loose material, taking care not to touch center of membrane. See Figure 15.

Figure 12

Figure 13

Figure 14

Figure 15

High Performance Ammonia Electrode User Guide

Note: Do not stretch the membrane.

11. Screw the membrane cap onto the outer body, being careful not to touch membrane. See Figure 16. Screw the membrane cap on half way and wrap any loose membrane material onto the threads and under the cap. Make sure the cap is fully screwed on. 12. Fill the outer body to the ll line with electrode lling solution. See Figure 17. 13. Insert the inner body into the outer body. Ensure that the inner body is fully inserted in the top of the outer body and then screw on the electrode cap. See Figure 18. 14. Gently tap the side of the electrode to remove air bubbles. 15. Soak the electrode in the ammonia electrode storage solution for at least 15 minutes before use. Prepare the storage solution by combining 100 mL of 1 ppm ammonia standard with 1 mL of alkaline reagent, Cat. No. 951011.

ll
ll

g lin n l lutio so
ll ll

Figure 16

Figure 17

Figure 18

High Performance Ammonia Electrode User Guide

Checking Electrode Operation (Slope)


This procedure measures electrode slope. Slope is dened as the change in millivolts observed with every tenfold change in concentration. Obtaining the slope value provides the best means for checking electrode operation. These are general instructions that can be used with most meters to check the electrode operation. See the meter user guide for more specic information. 1. If the electrode has been stored dry, prepare the electrode as described in the Electrode Preparation section.

2. Connect the electrode to the meter. 3. Place 100 mL of distilled water into a 150 mL beaker. Add 2 mL of pH-adjusting ISA, Cat. No. 951211. Stir the solution thoroughly. Set the meter to the mV mode. 4. Rinse the electrode by immersing it in distilled water, blot it dry and place it into the solution prepared in step 3. To prevent air entrapment on the membrane surface, be sure to use an electrode holder that keeps the electrode at a 20 angle. If air bubbles appear on the sensing membrane, gently tap the electrode to remove the bubbles. 5. Select either a 0.1 M ammonium chloride or 1000 ppm ammonia as nitrogen standard. Pipette 1 mL of the standard into the beaker and stir the solution thoroughly. When a stable reading is displayed, record the electrode potential in millivolts. 6. Pipette 10 mL of the same standard into the same beaker. Stir the solution thoroughly. When a stable reading is displayed, record the electrode potential in millivolts. 7 . There should be a 54 to 60 mV difference between the two millivolt readings when the solution temperature is between 20-25 C. If the millivolt potential is not within this range, refer to the Troubleshooting section. Note: If working with ammonia samples with concentrations of 1 ppm as nitrogen and lower, refer to the Low-Level Measurements by Direct Calibration section for a low-level checking electrode operation procedure.

High Performance Ammonia Electrode User Guide

Units of Measurement
Ammonia can be measured in units of moles per liter (M), parts per million as ammonia, parts per million as nitrogen or any other convenient unit. See Table 1. Table 1 Concentration Unit Conversion Factors

moles/liter (M) 10-4 10-3 10-2 10-1

ppm as nitrogen (N) 1.40 14.0 140 1400

ppm as ammonia (NH3) 1.70 17.0 170 1700

Sample Requirements
Samples must be aqueous; they must not contain organic solvents. Contact Technical Support for recommendations on the use of the electrode in unusual applications. Samples and standards should be at the same temperature. A 1 C difference in temperature will give rise to about 2% measurement error. In all analytical procedures, ISA must be added to all samples and standards immediately before measurement. After addition of the ISA, all solutions should fall within a pH 11 to 14 range (solutions should be blue in this range when using the pH-adjusting ISA, Cat. No. 951211) and have a total level of dissolved species below 1 M. If the total level of dissolved species is above 1 M, see the Effect of Dissolved Species section.

10

High Performance Ammonia Electrode User Guide

Measuring Hints
Store samples according to the Sample Storage section. Use beakers that minimize the ratio of surface area to volume. Keep beakers containing standards and samples covered between measurements. If working with samples that have a concentration of 1 ppm as nitrogen or higher, add 2 mL of pH-adjusting ISA (Cat. No. 951211) to every 100 mL of sample or standard. Once the ISA is added, a blue color should be observed. For samples that have a concentration of 1 ppm as nitrogen or lower, add 1 mL of alkaline reagent (Cat. No. 951011) to every 100 mL of standard or sample. ISA should be added immediately before a measurement is made to prevent ammonia loss. Between measurements, rinse the electrode by immersing it in distilled water. Be sure samples, standards and the electrode are at the same temperature. Samples and standards should be stirred using a magnetic stirrer or stirrer probe. Some magnetic stirrers generate enough heat to change solution temperature. Place a piece of insulating material such as cork or styrofoam between the beaker and the stirring plate to minimize this effect. Stir all standards and samples at a moderate, uniform rate. Verify calibration every two hours by placing the electrode in a fresh aliquot of the middle or high concentration standard used for calibration. If the value has changed, recalibrate the electrode. Always use fresh standards for calibration. After immersion in solution, check the electrode for any air bubbles on the membrane surface and remove bubbles by gently tapping the electrode. If electrode response is slow, replace the membrane and soak the electrode for at least 15 minutes in the ammonia electrode storage solution. Refer to the Electrode Preparation section.

High Performance Ammonia Electrode User Guide

11

Sample Storage
If possible, alkaline samples should be measured immediately. The rate of ammonia loss at room temperature from a stirred 100 mL basic solution in a 100 mL beaker is about 50% in six hours. If samples must be stored, make them slightly acidic (pH 6) by adding 0.5 mL of 1 M HCl to each liter of sample, and place samples in tightly capped vessels. Make stored samples basic by adding ISA immediately before measurement. For information on EPA regulations on ammonia sample storage, visit the Applications and Technical Resources section of our website at www.thermo.com/water.

Electrode Storage
Note: The membrane must not be allowed to dry out. If the membrane is left out of solution, replace the membrane. Refer to the Electrode Preparation section. Storage between measurements Store the electrode in the ammonia electrode storage solution. Prepare the storage solution by combining 100 mL of 1 ppm standard with 1 mL of alkaline reagent (Cat. No. 951011). To make the storage solution using the 0.1 M ammonia chloride standard, pipette 0.07 mL of the standard and 1 mL of the alkaline reagent into a 100 mL volumetric ask and dilute to the mark on the ask with distilled water. To make the storage solution using the 1000 ppm ammonia as nitrogen standard, pipette 0.1 mL of the standard and 1 mL of the alkaline reagent into a 100 mL volumetric ask and dilute to the mark on the ask with distilled water. To make the storage solution using the 100 ppm ammonia as nitrogen standard, pipette 1 mL of the standard and 1 mL of the alkaline reagent into a 100 mL volumetric ask and dilute to the mark on the ask with distilled water. Overnight storage and storage up to one week Store the electrode in the electrode lling solution, Cat. No. 951209. Storage over one week Disassemble the electrode and rinse all the components with distilled water. Dry and reassemble the electrode without electrode lling solution and without a membrane. When preparing the electrode for use, follow the procedures in the Electrode Preparation section.

12

High Performance Ammonia Electrode User Guide

Analytical Techniques
A variety of analytical techniques are available to the analyst. The following is a description of these techniques. Direct Calibration is a simple procedure for measuring a large number of samples. Only one meter reading is required for each sample. Calibration is performed using a series of standards. The concentration of the samples is determined by comparison to the standards. ISA is added to all solutions to ensure that samples and standards have similar ionic strength. Incremental Techniques provide a useful method for measuring samples, since calibration is not required. As in direct calibration, any convenient concentration unit can be used. The different incremental techniques are described below. They can be used to measure the total concentration of a specic ion in the presence of a large (50-100 times) excess of complexing agents. Known Addition is useful for measuring dilute samples, checking the results of direct calibration (when no complexing agents are present), or measuring the total concentration of an ion in the presence of an excess complexing agent. The electrode is immersed in the sample solution and an aliquot of a standard solution containing the measured species is added to the sample. From the change in potential before and after the addition, the original sample concentration is determined. Known Subtraction is useful as a quick version of a titration, or for measuring species for which stable standards do not exist. It is necessary to know the stoichiometric ratio between standard and sample. For known subtraction, an electrode sensing the sample species is used. Stable standards of a species reacting completely with the sample in a reaction of known stoichiometry are necessary.

High Performance Ammonia Electrode User Guide

13

Analate Addition is often used to measure soluble solid samples, viscous samples, small or very concentrated samples, to diminish the effects of complex sample matrices, or to diminish the effects of varying sample temperatures. This method is not suitable for dilute or low concentration samples. Total concentration is measured even in the presence of complexing agents. The electrode is immersed in a standard solution containing the ion to be measured and an aliquot of the sample is added to the standard. The original sample concentration is determined from the change in potential before and after the addition. Analate Subtraction is used in the measurement of ions for which no ion-selective electrode exists. The electrode is immersed in a reagent solution that contains a species that the electrode senses, and that reacts with the sample. It is useful when sample size is small, or samples for which a stable standard is difcult to prepare, and for viscous or very concentrated samples. The method is not suited for very dilute samples. It is also necessary to know the stoichiometric ratio between standard and sample.

Titrations are quantitative analytical techniques for measuring the concentration of a species by incremental addition of a reagent (titrant) that reacts with the sample species. Sensing electrodes can be used for determination of the titration end point. Ion selective electrodes are useful as end point detectors, because they are unaffected by sample color or turbidity. Titrations are approximately 10 times more precise than direct calibration, but are more time-consuming.

14

High Performance Ammonia Electrode User Guide

Typical Calibration Curve


In the direct calibration procedure, a calibration curve is constructed either in the meter memory or on semi-logarithmic paper. Electrode potentials of standard solutions are measured and plotted on the linear axis against their concentrations on the log axis. In the linear regions of the curves, only two standards are needed to determine a calibration curve. In non-linear regions, more points must be taken. Procedures are given in the Direct Calibration section for concentrations in the linear region of electrode response. The Low-Level Measurements By Direct Calibration section provides procedures for measurements in the non-linear region of electrode response.

-150 -100 -50 0 50 100 150 200 250

10-4

electrode potential (mV)

10-3

10-2

10-1

10

102

103

104

Log NH3-N Concentration (mg/L)

Figure 19: Typical Ammonia Electrode Calibration Curve

High Performance Ammonia Electrode User Guide

15

Direct Calibration
Setup 1. Prepare the electrode as described in the Electrode Preparation section. 2. Connect the electrode to the meter. 3. Prepare at least two standards that bracket the expected sample range and differ in concentration by a factor of ten. Standards can be prepared in any concentration unit to suit the particular analysis requirement. All standards should be at the same temperature as the samples. For details on temperature effects on electrode performance, refer to the Temperature Effects section. Serial Dilutions Serial dilution is the best method for the preparation of standards. Serial dilution means that an initial standard is diluted, using volumetric glassware, to prepare a second standard solution. The second standard is similarly diluted to prepare a third standard, and so on, until the desired range of standards has been prepared. 1. To prepare a 10-2 M NH4Cl standard (170 ppm as NH3 and 140 ppm as N) Pipette 10 mL of the 0.1 M NH4Cl standard into a 100 mL volumetric ask. Dilute to mark with deionized water and mix well. 2. To prepare a 10-3 M NH4Cl standard (17 ppm as NH3 and 14 ppm as N) Pipette 10 mL of the 10-2 M NH4Cl standard into a 100 mL volumetric ask. Dilute to mark with deionized water and mix well. 3. To prepare a 10-4 M NH4Cl standard (1.7 ppm as NH3 and 1.4 ppm as N) Pipette 10 mL of the 10-3 M NH4Cl standard into a 100 mL volumetric ask. Dilute to mark with deionized water and mix well. 4. To prepare a 10-5 M NH4Cl standard (0.17 ppm as NH3 and 0.14 ppm as N) Pipette 10 mL of the 10-4 M NH4Cl standard into a 100 mL volumetric ask. Dilute to mark with deionized water and mix well.

16

High Performance Ammonia Electrode User Guide

To prepare standards with a different concentration use the following formula: C1 * V1 = C2 * V2

where C1 = concentration of original standard V1 = volume of original standard C2 = concentration of standard after dilution V2 = volume of standard after dilution

For example, to prepare 100 mL of 25 ppm as NH3 standard from a 500 ppm as NH3 standard: C1 = 500 ppm as NH3 V1 = unknown C2 = 25 ppm as NH3 V2 = 100 mL 500 ppm * V1 = 25 ppm * 100 mL V1 = (25 ppm * 100 mL) / 500 ppm = 5 mL

To make the 25 ppm as NH3 standard: Pipette 5 mL of the 500 ppm as NH3 standard into a 100 mL volumetric ask. Dilute to mark with deionized water and mix well.

High Performance Ammonia Electrode User Guide

17

Direct Calibration Using a Meter with an ISE (Concentration) Mode Note: See the meter user guide for more specic information. 1. Measure 100 mL of the less concentrated standard into a 150 mL beaker. Add 2 mL of pH-adjusting ISA. Stir the solution thoroughly.

2. Rinse the electrode by immersing it in distilled water, blot it dry and place it into the beaker with the less concentrated standard. Wait for a stable reading and then adjust the meter to display the value of the standard, as described in the meter user guide. 3. Measure 100 mL of the more concentrated standard into a second 150 mL beaker. Add 2 mL of pH-adjusting ISA. Stir the solution thoroughly. 4. Rinse the electrode by immersing it in distilled water, blot it dry and place it into the beaker with the more concentrated standard. Wait for a stable reading and then adjust the meter to display the value of the second standard, as described in the meter user guide. 5. Record the resulting slope value. The slope should be between -54 to -60 mV when the standards are between 20-25 C. 6. Measure 100 mL of the sample into a 150 mL beaker. Add 2 mL of pH-adjusting ISA. Stir the solution thoroughly. 7. Rinse the electrode by immersing it in distilled water, blot it dry and place it into sample. The concentration will be displayed on the meter.

Note: Smaller volumes of solutions may be used as long as the ratio of standard or sample to pH-adjusting ISA does not change. For example, 50 mL of standard or sample requires the addition of 1 mL of pH-adjusting ISA.

18

High Performance Ammonia Electrode User Guide

Direct Calibration Using a Meter with a Millivolt Mode 1. Adjust the meter to measure mV.

2. Measure 100 mL of the less concentrated standard into a 150 mL beaker. Add 2 mL of pH-adjusting ISA. Stir the solution thoroughly. 3. Rinse the electrode by immersing it in distilled water, blot it dry and place it into the beaker with the less concentrated standard. When a stable reading is displayed, record the mV value and corresponding standard concentration. 4. Measure 100 mL of the more concentrated standard into a second 150 mL beaker. Add 2 mL of pH-adjusting ISA. Stir the solution thoroughly. 5. Rinse the electrode by immersing it in distilled water, blot it dry and place it into the beaker with the more concentrated standard. When a stable reading is displayed, record the mV value and corresponding standard concentration. 6. Using semi-logarithmic graph paper or a spreadsheet, prepare a calibration curve by plotting the millivolt values on the linear axis and the standard concentration values on the logarithmic axis. 7 . Measure 100 mL of the sample into a 150 mL beaker. Add 2 mL of pH-adjusting ISA. Stir the solution thoroughly.

8. Rinse the electrode by immersing it in distilled water, blot it dry and place it into the sample. When a stable reading is displayed, record the mV value. 9. Using the calibration curve prepared in step 6, determine the unknown concentration of the sample.

High Performance Ammonia Electrode User Guide

19

Low-Level Measurements By Direct Calibration


These procedures are for solutions with a concentration of 1 ppm ammonia as nitrogen (7 x 10-5 M, 1.2 ppm as ammonia) or less. For solutions low in ammonia but high in total ionic strength, perform the same procedure with one change: prepare a calibration solution with a background composition similar to the sample. Accurate measurement requires that the following conditions be met: Adequate time must be allowed for electrode stabilization. Longer response time will be needed for low-level measurements Remember to stir all standards and samples at a moderate, uniform rate Electrode response is relatively slow at low levels of ammonia and is faster going in the direction of increasing concentration. When measuring low levels of ammonia, keep samples and standards covered and work with large solution volumes to minimize surface-area-to-volume ratio. This avoids ammonia absorption from air. Time response is slow at ammonia concentrations of 0.1 ppm ammonia as nitrogen and lower. Allow up to 5 minutes for a stable reading in a low-level standard or sample. Setup 1. Prepare the electrode as described in the Electrode Preparation section. 2. Soak the electrode in ammonia electrode storage solution for at least 15 minutes. 3. Rinse the electrode by immersing it in distilled water and then blot it dry. 4. Connect the electrode to the meter.

20

High Performance Ammonia Electrode User Guide

Low-Level Checking Electrode Operation (Slope) This procedure measures electrode slope. Slope is dened as the change in millivolts observed with every tenfold change in concentration. Obtaining the slope value provides the best means for checking electrode operation. These are general instructions that can be used with most meters to check the electrode operation. See the meter user guide for more specic information. 1. Place 100 mL of distilled water into a 150 mL beaker. Add 1 mL of alkaline reagent, Cat. No. 951011. Stir the solution thoroughly. Set the meter to the mV mode.

2. Rinse the electrode by immersing it in distilled water, blot it dry and place it into the solution prepared in step 1. To prevent air entrapment on the membrane surface, be sure to use an electrode holder that keeps the electrode at a 20 angle. If air bubbles appear on the sensing membrane, gently tap the electrode to remove the bubbles. 3. Select either a 100 ppm ammonia as nitrogen standard or 10 ppm ammonia as nitrogen standard. Pipette 1 mL of the standard into the beaker and stir the solution thoroughly. When a stable reading is displayed, record the electrode potential in millivolts. 4. Pipette 10 mL of the same standard into the same beaker. Stir the solution thoroughly. When a stable reading is displayed, record the electrode potential in millivolts. 5. There should be a 54 to 60 mV difference between the two millivolt readings when the solution temperature is between 20-25 C. If the millivolt potential is not within this range, refer to the Troubleshooting section.

High Performance Ammonia Electrode User Guide

21

Low-Level Calibration 1. Prepare a 1 ppm ammonia as nitrogen standard by pipetting 10 mL of the 100 ppm ammonia as nitrogen standard into a 1 L volumetric ask and diluting to the mark with deionized water.

2. Prepare a 0.1 ppm ammonia as nitrogen standard by pipetting 50 mL of the 1 ppm ammonia as nitrogen standard into a 500 mL volumetric ask and diluting to the mark with deionized water. 3. Prepare a 0.01 ppm ammonia as nitrogen standard by pipetting 50 mL of the 0.1 ppm ammonia as nitrogen standard into a 500 mL volumetric ask and diluting to the mark with deionized water. 4. Measure 50 mL of each standard and pour the standards into separate, clean beakers. Just prior to calibration, add 0.5 mL of alkaline reagent, Cat. No. 951011, to each beaker. 5. Set the meter measurement mode to ISE. If using a Thermo Scientic Orion Star Series meter or equivalent meter, set the stirrer speed to 4 and the read type to continuous. In the ISE setup mode, set the resolution to 3, the units to mg/L, the range to low and turn on the blank correction feature. 6. Perform a three point calibration using the 0.01 ppm ammonia as nitrogen, 0.1 ppm ammonia as nitrogen and 1 ppm ammonia as nitrogen standards. Between standards, rinse the electrode by immersing it in distilled water and then blot it dry. Stir all solutions throughout the procedure. Calibrate daily using fresh solutions.

22

High Performance Ammonia Electrode User Guide

Measuring Ammonia in Solutions that Wet the Membrane


The membrane of the ammonia electrode is gas-permeable and hydrophobic. Liquid does not wet the membrane and liquid does not penetrate the membrane holes. If a sample solution is nonaqueous, or if it contains a surfactant that wets the membrane, the liquid will penetrate the membrane. This causes difculties in samples such as sewage, which contains surfactants, and samples that are nonaqueous, such as latex paint or nylon. To measure ammonia in such samples, the electrode should be suspended above the sample. If the ammonia electrode is placed in a closed system saturated with water vapor, the electrode reacts to ammonia in the gas phase. Measurements of solutions above 10-3 M ammonia (14 ppm ammonia as nitrogen) are possible under these conditions. To measure ammonia in samples containing surfactants, or nonaqueous solutions, adjust the sample pH to 11-13 with pH-adjusting ISA. Transfer the solution to a 125 mL Erlenmeyer ask containing a magnetic stirring bar. Fit the neck of the ask with a rubber stopper that has a hole large enough to hold the electrode snugly. The closed ask forms an airtight system whose gas phase is saturated with water vapor and has a partial pressure of ammonia in equilibrium with the solution. Normal analytical techniques may be used with the electrode in the gas phase. Calibrate the electrode in a closed ask of standards, or make a standard addition to the closed ask of sample. The electrode in the gas phase has a longer response time than if it were actually in a surfactant-free aqueous solution. Minimum air space between solution and electrode is necessary for best time response.

High Performance Ammonia Electrode User Guide

23

Measuring Organic Nitrogen


For information concerning the measurement of organic nitrogen, contact Technical Support. Within the United States call 1.800.225.1480 and outside the United States call 978.232.6000 or fax 978.232.6031. In Europe, the Middle East and Africa, contact your local authorized dealer. For the most current contact information, visit www.thermo.com/water. Nitrate Analysis Test Kit , Cat. No. 700005 The nitrate analysis test kit includes one 475 mL bottle each of alkaline reagent (Cat. No. 951011), reducing reagent (Cat. No. 700006), 100 ppm nitrate as nitrogen standard (Cat. No. 930707), 100 ppm ammonia as nitrogen standard (Cat. No. 951207); two 50 mL bottles of electrode lling solution (Cat. No. 951203); and application procedure number 115: Nitrate Measurement in Environmental Samples .

Application Notes
For application notes, such as measuring ammonia in wastewater, please visit www.thermo.com/water and select the applications link.

24

High Performance Ammonia Electrode User Guide

Known Addition
Known Addition is a convenient technique for measuring samples because no calibration curve is needed. It can be used to verify the results of a direct calibration or to measure the total concentration of an ion in the presence of a large excess of a complexing agent. The sample potential is measured before and after addition of a standard solution. Accurate measurement requires that the following conditions be met: Concentration should approximately double as a result of the addition. Sample concentration should be known to within a factor of three. In general, either no complexing agent or a large excess of the complexing agent must be present. Ratio of the uncomplexed ion to complexed ion must not be changed by addition of the standard. All samples and standards should be at the same temperature.

Setup 1. Prepare the electrode as described in the Electrode Preparation section. 2. Connect the electrode to the meter. 3. Prepare a standard solution that, upon addition to the sample, will cause concentration of the ammonia to double. Refer to Table 2 as a guideline. 4. Determine the slope of the electrode by performing the procedure in the Checking Electrode Operation (Slope) section. 5. Rinse the electrode between solutions by immersing it in distilled water and then blot it dry.

High Performance Ammonia Electrode User Guide

25

Using a Meter with a Known Addition Mode: Note: See the meter user guide for more specic information. 1. Set the meter to the known addition mode. 2. Measure 100 mL of the sample into a beaker. 3. Rinse the electrode by immersing it in distilled water, blot it dry and then place it in the sample solution. Add 2 mL of pH-adjusting ISA. Stir the solution thoroughly. 4. When a stable reading is displayed, set the meter as described in the meter user guide, if required. 5. Pipette the appropriate amount of the standard solution into the beaker. Stir the solution thoroughly. 6. When a stable reading is displayed, record the sample concentration. Using a Meter with a Millivolt Mode: 1. Set the meter to relative millivolt mode. 2. Measure 100 mL of the sample into a 150 mL beaker. Add 2 mL of pH-adjusting ISA. Stir the solution thoroughly. 3. Rinse the electrode by immersing it in distilled water, blot it dry and then place it in the sample solution. When a stable reading is displayed, set the millivolt reading to zero. If the reading cannot be set to zero, record the mV value. 4. Pipette the appropriate amount of standard solution into the beaker. Stir the solution thoroughly. 5. When a stable reading is displayed, record the mV value. If the meter could not be zeroed in step 3, subtract the rst reading from the second to nd E. 6. From Table 3, nd the Q value that corresponds to the change in potential, E. To determine the original sample concentration, multiply Q by the concentration of the added standard: Csam = QCstd Cstd = standard concentration Csam = sample concentration Q = reading from known addition table where:

26

High Performance Ammonia Electrode User Guide

The table of Q values is calculated for a 10% volume change for electrodes with slope in the range of -57 .2 to -60.2 mV. The equation for the calculation of Q for different slopes and volume changes is given below: p Q = (1 + p)10E/S - 1 Q S = reading from known addition table = slope of the electrode

where: E = E2 - E1 volume of standard p = volume of sample

Table 2 Volume of Addition 1 mL 5 mL 10 mL* Concentration of Standard 100 x sample concentration 20 x sample concentration 10 x sample concentration

* Most convenient volume to use

High Performance Ammonia Electrode User Guide

27

Table 3 Known Addition Table for an added volume one-tenth the sample volume. Slopes (in the column headings) are in units of mV/decade.
E Monovalent 5.0 5.2 5.4 5.6 5.8 6.0 6.2 6.4 6.6 6.8 7.0 7.2 7.4 7.6 7.8 8.0 8.2 8.4 8.6 8.8 9.0 9.2 9.4 9.6 9.8 10.0 10.2 10.4 10.6 10.8 11.0 11.2 11.4 11.6 11.8 12.0 12.2 12.4 12.6 12.8 13.0 13.2 13.4 13.6 13.8 14.0 14.2 14.4 14.6 14.8 15.0 15.5 16.0 16.5 17.0 (-57.2) 0.2894 0.2806 0.2722 0.2642 0.2567 0.2495 0.2436 0.2361 0.2298 0.2239 0.2181 0.2127 0.2074 0.2024 0.1975 0.1929 0.1884 0.1841 0.1800 0.1760 0.1722 0.1685 0.1649 0.1614 0.1581 0.1548 0.1517 0.1487 0.1458 0.1429 0.1402 0.1375 0.1349 0.1324 0.1299 0.1276 0.1253 0.1230 0.1208 0.1187 0.1167 0.1146 0.1127 0.1108 0.1089 0.1071 0.1053 0.1036 0.1019 0.1003 0.0987 0.0949 0.0913 0.0878 0.0846 Q Concentration Ratio (-58.2) 0.2933 0.2844 0.2760 0.2680 0.2604 0.2531 0.2462 0.2396 0.2333 0.2273 0.2215 0.2160 0.2107 02.056 0.2007 0.1961 0.1915 0.1872 0.1830 0.1790 0.1751 0.1714 0.1677 0.1642 0.1609 0.1576 0.1544 0.1514 0.1484 0.1455 0.1427 0.1400 0.1374 0.1349 0.1324 0.1300 0.1277 0.1254 0.1232 0.1210 0.1189 0.1169 0.1149 0.1130 0.1111 0.1093 0.1075 0.1057 0.1040 0.1024 0.1008 0.0969 0.0932 0.0897 0.0865 (-59.2) 0.2972 0.2883 0.2798 0.2717 0.2640 0.2567 0.2498 0.2431 0.2368 0.2307 0.2249 0.2193 0.2140 0.2088 0.2039 0.1992 0.1946 0.1902 0.1860 0.1820 0.1780 0.1742 0.1706 0.1671 0.1636 0.1603 0.1571 0.1540 0.1510 0.1481 0.1453 0.1426 0.1399 0.1373 0.1348 0.1324 0.1301 0.1278 0.1255 0.1233 0.1212 0.1192 0.1172 0.1152 0.1133 0.1114 0.1096 0.1079 0.1061 0.1045 0.1028 0.0989 0.0951 0.0916 0.0883 (-60.1) 0.3011 0.2921 0.2835 0.2754 0.2677 0.2603 0.2533 0.2466 0.2402 0.2341 0.2282 0.2226 0.2172 0.2120 0.2023 0.2023 0.1977 0.1933 0.1890 0.1849 0.1809 0.1771 0.1734 0.1698 0.1664 0.1631 0.1598 0.1567 0.1537 0.1507 0.1479 0.1451 0.1424 0.1398 0.1373 0.1348 0.1324 0.1301 0.1278 0.1256 0.1235 0.1214 0.1194 0.1174 0.1155 0.1136 0.1118 0.1100 0.1082 0.1065 0.1048 0.1009 0.0971 0.0935 0.0901

28

High Performance Ammonia Electrode User Guide

E Monovalent 17.5 18.0 18.5 19.0 19.5 20.0 20.5 21.0 21.5 22.0 22.5 23.0 23.5 24.0 24.5 25.0 25.5 26.0 26.5 27.0 27.5 28.0 28.5 29.0 29.5 30.0 31.0 32.0 33.0 34.0 35.0 36.0 37.0 38.0 39.0 40.0 41.0 42.0 43.0 44.0 45.0 46.0 47.0 48.0 49.0 50.0 51.0 52.0 53.0 54.0 55.0 56.0 57.0 58.0 59.0 60.0 (-57.2) 0.0815 0.0786 0.0759 0.0733 0.0708 0.0684 0.0661 0.0640 0.0619 0.0599 0.0580 0.0562 0.0545 0.0528 0.0512 0.0497 0.0482 0.0468 0.0455 0.0442 0.0429 0.0417 0.0405 0.0394 0.0383 0.0373 0.0353 0.0334 0.0317 0.0300 0.0285 0.0271 0.0257 0.0245 0.0233 0.0222 0.0211 0.0201 0.0192 0.0183 0.0174 0.0166 0.0159 0.0151 0.0145 0.0138 0.0132 0.0126 0.0120 0.0115 0.0110 0.0105 0.0101 0.0096 0.0092 0.0088

Q Concentration Ratio (-58.2) 0.0833 0.0804 0.0776 0.0749 0.0724 0.0700 0.0677 0.0655 0.0634 0.0614 0.0595 0.0576 0.0559 0.0542 0.0526 0.0510 0.0495 0.0481 0.0467 0.0454 0.0441 0.0428 0.0417 0.0405 0.0394 0.0383 0.0363 0.0344 0.0326 0.0310 0.0294 0.0280 0.0266 0.0253 0.0241 0.0229 0.0218 0.0208 0.0199 0.0189 0.0181 0.0172 0.0165 0.0157 0.0150 0.0144 0.0137 0.0131 0.0125 0.0120 0.0115 0.0110 0.0105 0.0101 0.0096 0.0092 (-59.2) 0.0852 0.0822 0.0793 0.0766 0.0740 0.0716 0.0693 0.0670 0.0649 0.0629 0.0609 0.0590 0.0573 0.0555 0.0539 0.0523 0.0508 0.0493 0.0479 0.0466 0.0453 0.0440 0.0428 0.0416 0.0405 0.0394 0.0373 0.0354 0.0336 0.0319 0.0303 0.0288 0.0274 0.0261 0.0249 0.0237 0.0226 0.0215 0.0205 0.0196 0.0187 0.0179 0.0171 0.0163 0.0156 0.0149 0.0143 0.0136 0.0131 0.0125 0.0120 0.0115 0.0110 0.0105 0.0101 0.0096 (-60.1) 0.0870 0.0839 0.0810 0.0783 0.0757 0.0732 0.0708 0.0686 0.0664 0.0643 0.0624 0.0605 0.0586 0.0569 0.055 0.0536 0.0521 0.0506 0.0491 0.0478 0.0464 0.0452 0.0439 0.0427 0.0416 0.0405 0.0384 0.0364 0.0346 0.0328 0.0312 0.0297 0.0283 0.0269 0.0257 0.0245 0.0233 0.0223 0.0212 0.0203 0.0194 0.0185 0.0177 0.0169 0.0162 0.0155| 0.0148 0.0142 0.0136 0.0130 0.0124 0.0119 0.0114 0.0109 0.0105 0.0101

High Performance Ammonia Electrode User Guide

29

30

High Performance Ammonia Electrode User Guide

Electrode Characteristics
Electrode Response
The electrode exhibits good time response (one minute or less) for ammonia concentrations above 1 ppm ammonia as nitrogen. Below this value, response times are longer and ammonia absorption from the air may become a source of error. Above 1.0 M, ammonia is rapidly lost to the air. Samples above 1.0 M ammonia concentration can be diluted before measurement. For solutions with low concentration in ammonia but high in total ionic strength, perform the same procedure with one change: prepare a calibration solution with a background composition similar to the sample. Accurate measurement requires that the following conditions be met. Adequate time must be allowed for electrode stabilization. Longer response time will be needed for low-level measurements. Remember to stir all standards and samples at a moderate, uniform rate. When plotted on semi-logarithmic paper, electrode potential response as a function of ammonia concentration is a straight line with a slope of about -58 mV per decade.

250

electrode potential (mV)

200

10 to 0.01 ppm NH3 as N 10 to 0.1 ppm NH3 as N

150 10 to 1 ppm NH3 as N

100

50

10 ppm NH3 as N time (sec) 50 100 150 200 250 300 10 to 100 ppm NH3 as N

-50

Figure 20: Typical Electrode Response To Step Changes in Ammonia Concentration

High Performance Ammonia Electrode User Guide

31

Temperature Effects
A change in temperature will cause electrode response to shift and change slope. Table 4 lists the variation of theoretical response with temperature. At 10-3 M, a 1 C temperature change gives rise to a 2% error. Samples and standards should be at the same temperature. Note that the higher the temperature, the faster the ammonia loss from solution. Table 4

Temperature (C) 0 5 10 15 20 25 30 35 40

Slope (mV) -54.20 -55.20 -56.18 -57 .17 -58.16 -59.16 -60.15 -61.14 -62.13

Interferences
Volatile amines interfere with electrode measurements. Most gases do not interfere as they are converted to the ionic form in basic solution. Ionic species cannot cross the gas-permeable membrane and are not direct electrode interferences. However, the level of ions in solution can change the solubility of ammonia. Standards and samples should have about the same level of ions in the solution and about the same level of dissolved species. Also, some metallic ions complex ammonia, causing falsely low results in direct measurements. The pHadjusting ISA, Cat. No. 951211, removes interferences from metallic ions.

32

High Performance Ammonia Electrode User Guide

pH Effects
The pH of all standards and samples must be adjusted above 11 before they can be measured.

Complexation
Ammonia forms metal complexes with a number of metal ions: mercury, silver, copper, gold, nickel, cobalt, cadmium, and zinc. Most of these metals are removed in the form of hydroxide complexes or precipitates in basic solution. When hydroxide is present at the 0.1 M level and the ammonia concentration is below 10-3 M, only mercury will appreciably complex ammonia. The total ammonia level of the sample will be measured if the mercury in the sample is preferentially bound to some other species. Iodide is recommended for this purpose, since it forms a soluble mercury complex at all pH levels. Use of pHadjusting ISA inhibits the formation of some these common metal complexes in the sample, because it contains a high concentration of hydroxide ion and a reagent that removes mercury and silver ions.

Effect of Dissolved Species


Water vapor is a potential electrode interference. Water can move across the membrane as water vapor, changing the concentration of the electrode lling solution under the membrane. Such changes will be seen as electrode drift. Water vapor transport across the membrane is not a problem if the total level of dissolved species in solution (osmotic strength) is below 1 M or the electrode and sample temperatures are the same. Addition of ISA to samples of low osmotic strength automatically adjusts them to the correct level. Samples with osmotic strengths above 1 M should be diluted before measurement. Dilution should not reduce the ammonia level below 10-5 M. Samples with high osmotic strengths (above 1 M) and low ammonium levels (below 10-5 M) can be measured without dilution if the osmotic strength of the electrode lling solution is adjusted. To adjust the electrode lling solution, add 4.25 g solid NaNO3 to each 100 mL electrode lling solution.

Membrane Life
A membrane will last from one week to several months, depending on usage. Membrane failure is characterized by a shift in electrode potential, drift or poor response. Refer to Troubleshooting section.

High Performance Ammonia Electrode User Guide

33

Theory of Operation
The ammonia electrode uses a hydrophobic gas-permeable membrane to separate the sample solution from the electrode lling solution. Dissolved ammonia in the sample solution diffuses through the membrane until the partial pressure of ammonia is the same on both sides of the membrane. In any given sample the partial pressure of ammonia will be proportional to its concentration. Ammonia diffusing through the membrane dissolves in the lling solution and, to a small extent, reacts reversibly with water in the lling solution. NH3 + H2O NH4+ +OH-

The relationship between ammonia, ammonium ion and hydroxide is given by the following equation: [NH4] [OH-] [NH3]

= constant

The electrode lling solution contains ammonium chloride at a sufciently high level so that the ammonium ion concentration can be considered xed. Thus: [OH-] = [NH3] constant

The potential of the electrode sensing element with respect to the internal reference element is described by the Nernst equation: E = E0 - S log [OH-]

where: E = measured electrode potential E0 = reference potential OH- = hydroxide concentration in solution S = electrode slope (-59.2 mV/decade)

Since the hydroxide concentration is proportional to ammonia concentration, electrode response to ammonia is also Nernstian. E = E0 - S log [NH3]

The reference potential, E0, is partly determined by the internal reference element that responds to the xed level of chloride in the lling solution.

34

High Performance Ammonia Electrode User Guide

Ammonium Ion
When ammonia is dissolved in water it reacts with hydrogen ion to form ammonium ion: NH3 + H3O+ NH4+ + H2O

The relative amount of ammonia and ammonium ion is determined by the solution pH (see Figure 21). In acid solution, where hydrogen ion is readily available, virtually all the ammonia is converted to ammonium ion. At a pH of about 9.3, half the ammonia will be in the form of ammonium ion. Theoretically, it is possible to calculate the ratio of ammonia to ammonium ion, if the pH is known. The equilibrium constant for the reaction is: [NH4+] [H3O+][NH3] = [NH4+] 10-pH [NH3] = K 10-9.3

at 25 C = 0.1 and pK = 9.3. The ratio of ammonium to ammonia is given by: [NH4+] [NH3]

= K 10-pH = 109.3-pH

Martell, A.; Smith, R., Critical Stability Constants, Plenum Press. New York, NY, 1974. The exact value of K will vary with both temperature and ionic strength. For example, while the pK, at 25 C and = 0.1, is given as 9.3 (as in the discussion above) an increase in ionic strength to = 1.0 yields a pK of 9.4.

High Performance Ammonia Electrode User Guide

35

Partial Pressure of Ammonia


As discussed in the Theory of Operation section, the ammonia electrode responds to the partial pressure of dissolved ammonia gas. The partial pressure of dissolved ammonia gas is related to the ammonia concentration by Henrys Law: Kh = [NH3] aqueous PNH3 The Henrys Law constant, Kh, varies both with temperature and the level of dissolved species. For example, the constant is about 20% lower in 1 M NaCl that in distilled water. To keep the Henrys Law constant close to the same value, standards and samples should contain the same level of dissolved species and be about the same temperature. = 56 moles/liter - atm. (25 C)

% of species ammonia 100 80 60 40 20 ammonium 0 6 7 solution pH 8 9 10 11 12 13 14

Figure 21: Percent of Ammonia and Ammonium as a Function of pH

36

High Performance Ammonia Electrode User Guide

Troubleshooting
Troubleshooting Checklist
Symptom: Off-scale or over-range reading Membrane failure Replace the membrane. Refer to the Electrode Preparation section for instructions. Inner body not properly conditioned Soak the inner body in electrode lling solution for at least two hours. For best results, soak the inner body overnight. Inner body defective Refer to the Troubleshooting Guide section and perform the checking inner body procedure. Electrode lling solution not added Fill the electrode up to the ll line with electrode lling solution. Air bubble on the membrane Remove bubbles by gently tapping the side of the electrode. Electrode not in solution Insert the electrode in solution. Electrode not plugged into the meter properly Unplug and reconnect the electrode. Defective meter Perform the meter checkout procedure (refer to the meter user guide). Symptom: Low slope or no slope Membrane failure Replace the membrane. Refer to the Electrode Preparation section for instructions. Inner body defective Refer to the Troubleshooting Guide section and perform the checking inner body procedure. Standards contaminated or made incorrectly Prepare fresh standards. ISA not used or incorrect ISA used Use pH-adjusting ISA, Cat. No. 951211, when working with ammonia samples that have concentrations of 1 ppm ammonia as nitrogen or higher. Add 2 mL of pH-adjusting ISA to every 100 mL of standard and sample. Use alkaline reagent (low-level ISA), Cat. No. 951011, when working with ammonia samples that have concentrations of 1 ppm ammonia as nitrogen or lower. Add 1 mL of alkaline reagent to every 100 mL of standard and sample. Electrode exposed to air for extended period Replace the membrane. Refer to the Electrode Preparation section for instructions.

High Performance Ammonia Electrode User Guide

37

Symptom: Noisy or unstable readings (erratic, rapidly changing) Insufcient electrode lling solution Fill the electrode up to the ll line with electrode lling solution. Membrane cap loose Ensure that the membrane cap is screwed on tight enough to close the gap between the cap and body. Inner body defective Refer to the Troubleshooting Guide section and perform the checking inner body procedure. ISA not used Use the recommended ISA. Defective meter Perform the meter checkout procedure (refer to the meter user guide). Symptom: Wrong answer but calibration curve is correct Standards contaminated or made incorrectly Prepare fresh standards. Incorrect scaling of semi-logarithmic paper Refer to the Direct Calibration section. Incorrect millivolt sign used Make sure to correctly record the sign of mV values. Incorrect units used Apply the correct conversion factor: 10-3 M = 17 ppm as NH3 = 14 ppm as N Complexing agents in sample Use known addition, titration techniques or a decomplexing procedure. ISA added to standards, but not samples Add the same proportion of ISA to all standards and samples. Both ISA solutions used for one analysis To keep the dilution ratio the same, use only one type of ISA for each calibration and measurement set. Use pH-adjusting ISA, Cat. No. 951211, with ammonia samples that are 1 ppm and higher. Use alkaline reagent (low-level ISA), Cat. No. 951011, with ammonia samples that are 1 ppm and lower.

38

High Performance Ammonia Electrode User Guide

Symptom: Drift (reading slowly changing in one direction) Membrane failure Replace the membrane. Refer to the Electrode Preparation section for instructions. Inner body defective Refer to the Troubleshooting Guide section and perform the checking inner body procedure. Ammonia loss from sample sitting too long Reduce the surface area to volume ratio, slow the rate of stirring and avoid high temperatures. Filling solution leaking Ensure that the membrane is installed properly. Refer to the Electrode Preparation section. Electrode not assembled properly Ensure that the inner body is fully inserted in the top of the outer body when assembling the electrode. Refer to the Electrode Preparation section. Samples and standards at different temperatures Allow solutions to come to room temperature before measurement. Incorrect electrode lling solution used Fill the electrode using the correct electrode lling solution, Cat. No. 951209. Total level of dissolved species above 1 M Dilute solutions. Meter or stirrer improperly grounded Check meter and stirrer for grounding issues. Solutions not at constant temperature Allow solutions to come to room temperature before use. Magnetic stirrer generating heat Place insulating material between the magnetic stirrer and beaker. Electrode exposed to air for extended period Replace the membrane. Refer to the Electrode Preparation section for instructions.

High Performance Ammonia Electrode User Guide

39

Troubleshooting Guide
The most important principle in troubleshooting is to isolate the components of the system and check each in turn. The components of the system are the meter, electrode, standard, sample and technique. Meter The meter is the easiest component to eliminate as a possible cause of error. Thermo Scientic Orion meters include an instrument checkout procedure in the meter user guide and a shorting strap for convenience in troubleshooting. Consult the user guide for complete instructions and verify that the instrument operates as indicated and is stable in all steps. Electrode 1. Rinse the electrode thoroughly with distilled water.

2. Perform Checking Electrode Operation (Slope) procedure. 3. If the electrode fails this procedure, replace the membrane and soak the electrode as directed in the Electrode Preparation section. 4. Repeat the Checking Electrode Operation (Slope) procedure. 5. If the electrode still does not perform correctly, perform the Checking Inner Body procedure to determine if the inner body is working properly. Checking Inner Body: Solution 1 pH 4.01 buffer with 0.1 M NH4Cl or 0.1 M NaCl

 Take 100 mL of pH 4.01 buffer, Cat. No. 910104, and add 0.54 g of reagent-grade NH4Cl or 0.58 g of reagent-grade NaCl. Thoroughly stir the solution and label the bottle as Solution 1. Store the buffer for repeated use. Discard the buffer if turbidity develops. Solution 2 pH 7 .00 Buffer with 0.1 M NH4Cl or 0.1 M NaCl

 Take 100 mL of pH 7 .00 buffer, Cat. No. 910107, and add 0.54 g of reagent-grade NH4Cl or 0.58 g of reagent-grade NaCl. Thoroughly stir the solution and label the bottle as Solution 2. Store the buffer for repeated use. Discard the buffer if turbidity develops.

40

High Performance Ammonia Electrode User Guide

Note: The temperature of the buffers and distilled water must be 25 C 4 C and all solutions should be at same temperature within 1 C. Disassemble the ammonia electrode. If the electrode is dry, soak the glass tip of the inner body in lling solution for at least two hours. Rinse the inner body with distilled water and immerse it in pH 7 buffer with 0.1 M NH4Cl added. Make sure that the coiled reference wire is completely covered. Stir the solution throughout the procedure. Record the millivolt reading after two minutes. Rinse the inner body in distilled water and place in the pH 4 buffer with 0.1 M NH4Cl added. Watch the change in meter reading carefully. The reading should change by 100 mV in less than 30 seconds after immersion in the pH 4 buffer. After three minutes the mV difference between pH 7 and pH 4 should be greater than 150 mV if the inner body is operating correctly.

6. If the stability and slope check out properly, but measurement problems still persist, the sample may contain interferences or complexing agents, or the technique may be in error. Refer to the Standard, Sample, and Technique sections. 7 . Before replacing a faulty electrode, review the instruction manual and be sure to: Clean the electrode thoroughly Prepare the electrode properly  Use proper electrode lling solution, ISA and standards Review the Troubleshooting Checklist section

Standard The quality of results depends greatly upon the quality of the standards. Always prepare fresh standards when problems arise, it could save hours of frustrating troubleshooting. Errors may result from contamination of prepared standards, accuracy of dilution, quality of distilled water, or a mathematical error in calculating the concentrations. The best method for preparation of standards is by serial dilution. Refer to the Direct Calibration section.

High Performance Ammonia Electrode User Guide

41

Sample If the electrode works properly in standards but not in the sample, look for possible interferences, complexing agents or substances that could alter the electrode response or physically damage the electrode. If possible, determine the composition of the samples and check for problems. Refer to the Sample Requirements, Interferences, and pH Requirements sections. Technique Check the method of analysis for compatibility with your sample. Direct measurement may not always be the method of choice. If large amounts of complexing agents are present, known addition may be best method. If the sample is viscous, analate addition may solve the problem. If working at low levels, be sure to follow the low-level measurement technique. Also, be sure that the expected concentration of the ion of interest is within the electrodes limit of detection. If problems persist, review operational procedures and instruction manuals to be sure that proper technique has been followed. Refer to the Measuring Hints and Analytical Techniques sections.

Assistance
After troubleshooting all components of your measurement system, contact Technical Support. Within the United States call 1.800.225.1480 and outside the United States call 978.232.6000 or fax 978.232.6031. In Europe, the Middle East and Africa, contact your local authorized dealer. For the most current contact information, visit www.thermo.com/water.

Warranty
For the most current warranty information, visit www.thermo.com/water.

42

High Performance Ammonia Electrode User Guide

Ordering Information
Cat. No. 9512HPBNWP 9512HP01 951214* 951215* 951211 951011 951209 951006 951007 951207 Description High performance ammonia electrode with waterproof BNC connector High performance ammonia electrode with U.S. standard connector Loose membranes, box of 20 Preassembled outer body and membrane cap assembly, 3 pack pH-adjusting ISA, 475 mL bottle Alkaline reagent (low-level ISA), 475 mL bottle Electrode lling solution, 60 mL bottle 0.1 M NH4Cl standard, 475 mL bottle 1000 ppm ammonia as nitrogen standard, 475 mL bottle 100 ppm ammonia as nitrogen standard, 475 mL bottle

*  Use

with the 9512HPBNWP and 9512HP01 high performance ammonia electrodes only

High Performance Ammonia Electrode User Guide

43

Specications
Concentration Range
5 x 10-7 to 1 M 0.01 to 17 ,000 ppm as ammonia (NH3) 0.01 to 14,000 ppm as nitrogen (N)

pH Range
Samples and standards must be adjusted above pH 11

Temperature Range
0 to 50 C

Electrode Resistance
Less than 1,500 megohms

Slope
-54 to -60 mV (in the linear range of the electrode, with all solutions at 25 C 4 C and with all solutions at same temperature within 1 C)

Response Time
1 minute or less (at higher concentrations and with ideal electrode conditions)

Size
Electrode Length: 115 mm Body Diameter: 12 mm Cap Diameter: 16 mm Cable Length: 1 meter

Note: Specications subject to change without notice

44