Kerr Micros
Kerr Micros
Kerr Micros
Kerr microscopy
on a ferrimagnetic garnet
Fabian Lux
Universit Pierre et Marie Curie
M2 Nanomat
Abstract
This laboratory course gives an introduction to the magneto-optical Kerr effect in application to microscopic imaging of magnetic materials. In particular the ferrimagnetic hysteresis of GdPrBi3 (FeAl)5 O12
grown on a Gd3 Ga5 O12 garnet is measured. Besides, the macroscopic Faraday effect is observed.
Contents
I
Introduction
A
The Setup . . . . . . . . . . . . .
B
The Theory . . . . . . . . . . . . .
1
1
1
II Faraday Effect
3
3
5
6
I.
A.
Introduction
The Setup
B.
The Theory
(1)
Halogen lamp
Interference filter
l = 600 nm
Aperture diaphragm
Polarizer / Analyzer
Analyzer
BFP
Coil
CCD
Sample holder
Beam splitter
Computer
exx ( )
e = exy (, M )
0
exy (, M )
eyy ( )
0
0
0 .
ezz ( )
(n+ n )
Im(exy (, M))
2c
2cn
II.
Faraday Effect
The first part of this laboratory course investigates the Faraday effect in transmission. To
measure the effect, the light is polarized before the beam splitter and analyzed behind the
sample holder in two configurations. The first
without any sample and the second with transmission through the garnet (for zero and nonzero magnetic field). To obtain some statistics
each of these measurements is repeated three
times. The applied field strength is 100 Oe,
corresponding to a coil current of 2 A as controlled by the LabView program. Due to this
high current, the available time for the experiment is limited.
can be seen from the lecture notes, also a change of ellipticity results.
Figure 2: Image processing for the calibration of the microscope. First, a well focused area is selected. The contrast in
this area is enhanced and a two dimensional Fourier transformation is performed.
Incoming
260 550
260 450
260 550
260 520 60
100 Oe
271 180
271 220
271 500
271 300 170
+100 Oe
251 00
251 50
251 100
251 50 50
dL
100 Oe
10 380 180
(1 460 30 )/m
+100 Oe
9 470 80
(1 380 10 )/m
III.
A.
Kerr Effect
Calibration
Since the setup is an infinity system the theoretical magnification Mth is given by
Mth =
fOc
= 20.
fOb
(2)
= 0.915 m.
(3)
2N A
In principle the magnification can be increased
(by increasing the focal length of the ocular)
until the Rayleigh criterium makes it pointless
to magnify further. To improve the resolution
and achieve a better magnification, an objective with a larger aperture could be chosen.
For Mth = 20, the spacing of two pixels corresponds to the distance
R = 1.22
6.45
1
m = 0.3225 m R.
(4)
20
3
Therefore, its not useful to increase the magnification.
To determine the real magnification the
Kerr microscope is focused on the calibration
sample. The ambient light is turned off and
the sample is adjusted manually to the focal
point. To improve further, piezoelectric crystals
are used to move the sample. Then the device
is focused on the calibration array described
above. An image of this array is saved with
the computer software. The following procedure of image processing is illustrated in figure
2. First, a well focused area is selected. The
contrast in this area is enhanced by applying
a high pass filter with a certain cutoff. Then,
a two dimensional Fourier transformation is
performed. Fourier amplitudes in x and y directions are shown in figure 3 for the lowest
frequencies.
Kexp =
Mexp =
(7)
dCCD
Kexp = 24.3 0.3.
pixel
(8)
0.8
pixels- 1
Rel. Amplitude
1
pixel
= (3.77 0.04)
f dCCD
m
c2 / ndf = 0.03
0.06
0.4
(6)
1.0
0.6
(5)
f = 0.0106
0.0001
0.04
0.02
0.2
0.0
0.00
0.00
0.01
0.02
0.03
0.04
0.05
0.06
pixels-1
Order
B.
Domain width
(9)
1
= (19 3) m
f Kexp
(10)
P
d garnet
1
= 3.2 0.5
f Kexp
20 MS4 d2 2C
J
= (6.5 2.5) 1012
16Ku
m
0.04
0.03
0.02
0.01
(11)
0.00
0.00
0.01
0.02
0.03
0.04
0.05
pixel-1
Figure 7: Hysteresis loop for the garnet sample. Total intensity values are calculated by averaging the intensity of each
image.
C.
Hysteresis
In the last part a hysteresis loop of the garnet is observed with the Kerr microscope. A
LabView program controls to coil current and
triggers the CCD camera to take images at specific stages of the applied magnetic field, which
is varied between 100 Oe and +100 Oe. To
obtain the hysteresis curve, the intensity is averaged for each each recorded image. The average intensity decodes the total magnetization
of the sample. It is therefore no surprise, that
the recorded curve as shown in figure 7 is similar to the one recorded by a SQUID device.
Also, figure 7 shows the associated images as
obtained by the CCD camera. This gives a microscopic interpretation of the hysteresis loop.
In the process of decreasing the absolute
field the system stays remanent up to a certain point. At this critical field, domains with
1.0
0.8
0.6
0.4
0.2
0.0
Amplitude
0.6
0.4
0.2
0.0
0
0.6
0.4
0.2
0.0
1
2
20
40
60
Frequency [1/pixel]
5
-3
80x10