Hach 8009 Zinc Ed 08
Hach 8009 Zinc Ed 08
Hach 8009 Zinc Ed 08
DOC316.53.01145
Method 8009
Powder Pillows
Scope and application: For water and wastewater. Digestion is required for a total zinc analysis.
1
2
USEPA approved for wastewater analyses 3500 Zn B: Federal Register, 45(105) 36166 (May 29, 1980).
Adapted from Standard Methods for the Examination of Water and Wastewater.
Test preparation
Instrument-specific information
Table 1 shows all of the instruments that have the program for this test. The table also
shows sample cell and orientation requirements for reagent addition tests, such as
powder pillow or bulk reagent tests.
To use the table, select an instrument, then read across to find the applicable information
for this test.
Table 1 Instrument-specific information
Instrument
Sample cell
DR 6000
2495402
DR 3800
DR 2800
DR 2700
DR 1900
DR 5000
DR 3900
DR 900
2401906
Before starting
Install the instrument cap on the DR 900 cell holder before ZERO or READ is pushed.
Clean all glassware with 6.0 N (1:1) hydrochloric acid, then fully rinse with deionized water to remove contaminants.
Use only glass-stoppered mixing cylinders in this procedure.
Make sure that the dropper that is used in this procedure is plastic. Droppers that have rubber bulbs can contaminate the
reagent.
The reagents that are used in this test contain potassium cyanide. Keep cyanide solutions at pH > 11 to prevent
exposure to hydrogen cyanide gas. Collect the reacted samples for proper disposal.
Review the Safety Data Sheets (MSDS/SDS) for the chemicals that are used. Use the recommended personal protective
equipment.
Do not use the Pour-Thru Cell or sipper module (for applicable instruments) with this test.
Dispose of reacted solutions according to local, state and federal regulations. Refer to the Safety Data Sheets for disposal
information for unused reagents. Refer to the environmental, health and safety staff for your facility and/or local regulatory
agencies for further disposal information.
Items to collect
Description
Quantity
Cyclohexanone
0.5 mL
Sample cells (For information about sample cells, adapters or light shields, refer to Instrumentspecific information on page 1.)
Collect samples in clean glass or plastic bottles that have been cleaned with 6 N (1:1)
hydrochloric acid and rinsed with deionized water.
To preserve samples for later analysis, adjust the sample pH to less than 2 with
concentrated nitric acid (about 2 mL per liter). No acid addition is necessary if the
sample is tested immediately.
Keep the preserved samples at room temperature for a maximum of 6 months.
Before analysis, adjust the pH to 45 with 5 N sodium hydroxide solution. Do not
exceed pH 5 as zinc can precipitate. Do not add the reagent to samples with pH less
than 4 because there is cyanide in the reagent.
Correct the test result for the dilution caused by the volume additions.
Start
Zero
Read
Interferences
Interfering substance
Interference level
Aluminum
Cadmium
Copper
Iron (ferric)
Manganese
Interfering substance
Interference level
Nickel
Organic Material
Can prevent the correct pH adjustment of the sample by the reagents. Sample pretreatment may be necessary. Adjust the pH to 45.
Amino-tri(methylene phosphonic
acid) (AMP)
Samples that contain AMP cause a negative interference. Digest the sample to remove
this interference (use the total phosphorus hot plate digestion, Method 8190).
Note: Be sure to adjust the pH of the sample after the digestion to pH 45 with sodium hydroxide
before the zinc analysis.
Digestion
For total zinc determinations, the sample must be digested with heat and acid to make
sure that all forms of the metal are measured. The steps that follow can be used for a
mild digestion.
Note: The following procedure is the USEPA mild digestion procedure. Refer to the Water Analysis
Guide for more digestion procedures.
1. Add concentrated nitric acid to the sample with a glass serological pipet and pipet
filler:
2.
3.
4.
5.
6.
7.
If the sample was acidified for preservation, add 3 mL of nitric acid to 1 liter of the
preserved sample.
If the sample was not acidified for preservation, add 5 mL of nitric acid to 1 liter of
sample.
Transfer 100 mL of acidified sample to a 250-mL Erlenmeyer flask.
Add 5 mL of 1:1 hydrochloric acid.
Put the sample on a hot plate at 95 C (203 F) until 1520 mL of the sample
remains. Make sure that the sample does not boil.
Put the cooled sample through a 0.45-m filter to remove any insoluble material.
Adjust the pH of the digested sample to pH 45 with 5.0 N sodium hydroxide. Do not
exceed pH 5 as zinc may precipitate.
Quantitatively transfer the sample to a 100-mL volumetric flask and dilute to the mark
with deionized water.
Accuracy check
Standard additions method (sample spike)
Use the standard additions method (for applicable instruments) to validate the test
procedure, reagents and instrument and to find if there is an interference in the sample.
Items to collect:
1. Use the test procedure to measure the concentration of the sample, then keep the
(unspiked) sample in the instrument.
2. Go to the Standard Additions option in the instrument menu.
3. Select the values for standard concentration, sample volume and spike volumes.
4. Open the standard solution.
5. Prepare three spiked samples: use the TenSette pipet to add 0.1 mL, 0.2 mL and
0.3 mL of the standard solution, respectively, to three 20-mL portions of fresh sample.
Mix well.
6. Use the test procedure to measure the concentration of each of the spiked samples.
Start with the smallest sample spike. Measure each of the spiked samples in the
instrument.
7. Select Graph to compare the expected results to the actual results.
Note: If the actual results are significantly different from the expected results, make sure that
the sample volumes and sample spikes are measured accurately. The sample volumes and
sample spikes that are used should agree with the selections in the standard additions menu. If
the results are not within acceptable limits, the sample may contain an interference.
Method performance
The method performance data that follows was derived from laboratory tests that were
measured on a spectrophotometer during ideal test conditions. Users can get different
results under different test conditions.
Program
Standard
Sensitivity
Concentration change per 0.010 Abs change
780
1.00 mg/L Zn
0.971.03 mg/L Zn
0.013 mg/L Zn
Summary of method
Zinc and other metals in the sample are complexed with cyanide. Adding cyclohexanone
causes a selective release of zinc. The zinc reacts with 2-carboxy-2'-hydroxy-5'sulfoformazyl benzene (zincon) indicator to form a blue-colored species. The blue color is
masked by the brown color from the excess indicator. The intensity of the blue color is
proportional to the amount of zinc present. The measurement wavelength is 620 nm for
spectrophotometers or 610 nm for colorimeters.
Quantity/test
Unit
Item no.
100 tests
2429300
0.5 mL
100 mL MDB
1403332
100/pkg
2106669
Quantity/test
Unit
Item no.
each
2088640
Description
Unit
Item no.
Water, deionized
4L
27256
100 mL
237842
16/pkg
1424610
100 mL
1417742
Description
Unit
Item no.
each
50546
each
1206701
500 mL
88449
500 mL
15249
each
1970001
Required apparatus
Description
Mixing cylinder, graduated, 25-mL with stopper
Recommended standards
50/pkg
2185696
each
2196800
each
1451538
each
1465100
each
1457453
HACH COMPANY
WORLD HEADQUARTERS
Telephone: (970) 669-3050
FAX: (970) 669-2932
01/2014, Edition 8