A12 (E) User Manual
A12 (E) User Manual
A12 (E) User Manual
XK3190-A12
User Manual
A12-51007 V1.0
Content
CHAPTER 1 MAIN SPECIFICATION .......................-2CHAPTER 2 INSTALLATION .......................................-32.1 FRONT AND BACK VIEW OF INDICATOR ....-32.2 KEY FUNCTIONS .......................................-42.3 CONNECTING LOAD CELL TO INDICATOR.. ....................-4CHAPTER 3 OPERATION .......-53.1 POWER ON & ZERO SETTING....... -53.2 MANUAL ZERO SETTING.........-53.3 TARE .....-53.4 ACCUMULATING ....-53.5 COUNTING ......-53.6 USER FUNCTION STEETING ..-63.7 CONNECT SCOREBOARD TO INDICATOR .....-63.8 SERIAL COMMUNICATION AND INDICATOR COMMECTION ..-8CHAPTER 4 CALIBRATION .......................-9CHAPTER 5 ERROR INDICATION .. .. ....................-10CHAPTER 6 CHARGEABLE BATTERY .-10CHAPTER 7 MAINTENTANCE ..........................-11-
A12-51007 V1.0
-1-
2. Accuracy:
Grade, n=3000
3. Sample Rate
10 times / second
1.53mV / V
5. Scale interval
6. Display
Battery DC6V/4AH
04090%RH
-2050
A12-51007 V1.0
-2-
Chapter 2 Installation
2.1 FRONT AND BACK VIEW OF THE INDICATOR
Front View
Back View
A12-51007 V1.0
-3-
[FUNC ]
operator setting mode; less than 5 seconds, it will come into counting mode.
[ACCUM ]
[TARE ]
[ZERO ]
[ON/OFF ]
Press this button to start the indicator when it is off; and press it to shut off upon on.
The 9-pin socket is used for the link-up of load cell, which has been clearly shown in the graph
2-3.
2.
The 4-pin shielded cable is used, and the indicator does not have the function of long distance
compensation.
3.
Indicator must be reliably connected to Load cell and shielded-cable of load cell must be
reliably connect ed to underground. If indicator is powered on, the user should not insert or
withdraw the plug in order to protect the indicator and load cell.
4.
Sensor and indicator are static sensitive devices; you must adopt anti-static measures. The
electric wielding operation and other strong electric operation are prohibited. In order to protect
the operator, indicator, and relevant devices, you should install lightning rod in the
thunderstorm frequently happening area
Port of Indicator
-V
Shielded
+V
-IN
+IN
(Graph 2-3)
A12-51007 V1.0
+V
-V
+IN
-IN
shielded
-4-
Output Signal+
Output Signal -
Chapter 3 Operation
3.1 POWER ON AND AUTO ZERO-SETTING
3.1.1 The indicator will perform 999999-000000 to self-checking when turning on. Then it will enter
weighing mode.
3.1.2 When power on, if loading weight on the scale deviates from the zero point, but still within zero
set range, the indicator will set zero automatically; if out of range, it is necessary to adjust the
zero point or recalibrate or reset.
-5-
X=1:
X=2:
x
X=1:
X=2:
No this function
Power off 10 minutes later
20 minutes
X=4
30 minutes
3P3
X=1:
X=2:
X=3:
X=4:
x
X=1:
X=2:
X=3:
X=4:
X=5:
X=6:
KeepPrinter
x
X=1:
X=2:
X=3:
7P7
X=1:
X=2:
6P6
X=3
4P4
Lb change
kg display
Lb display
2P2
5P5
kg
A12-51007 V1.0
Backlight setting
No backlight
Automatic backlight
Keep lighting
Zero-tracking scope
-6-
X=1:
X=2:
X=3:
X=4:
X=5:
X=6:
X=7:
8P8
0.5e
1.0e
1.5e
2.0e
2.5e
3.0e
5.0e
X=1:
X=2:
X=3:
X=4:
9P9
2%FS
4%FS
10%FS
20%FS
X=1:
X=2:
X=3:
X=4:
10P10
2%FS
4%FS
10%FS
20%FS
x
X=1:
X=2:
X=3:
11. P11
X=1:
X=2:
X=3
12. P12
X=1:
X=2:
X=3:
Stabile time
high
middle
low
Stable extent
low
middle
high
with connection, damage will happen to output port of instrument and input port of scoreboard,
sometimes, the damage is so big to influence the instrument and scoreboard. Only specially
provided connecting cable is allowed to be used.
A12-51007 V1.0
-7-
Make sure that communication interface output lead and computer are correctly connected, if
there is something wrong with connection, damage will happen to output port of instrument and
input port of computer, sometimes, the damage is so big that instrument, computer and
corresponding peripherals are got involved.
Necessary computer technology and programming expertise are required for computer
communication, which should be participated and instructed by professionals. Non-professional
staff is supposed not to be involved in this regard.
With RS232(optional) serial communication interface, the XK3190-A12 indicator can be connected to
computer for communication.
1. All data are ASCII code, every set of which is composed of 10 bits: the 1st is starting bit, the 10th is
stop bit, the middle in between are 8 data bits.
Communication mode as follows:
(1). In continuous mode:
The data transmitted is weight (Gross weight or net weight)
The format of G.W.ww000.000kg or ww000.000lb
The format of N.W wn000.000kg or wn000.000lb
NoteThe position of above decimal is decided by the decimal set on the indicator.
(2). In command mode:
The indicator performs the corresponding operation according to the command transmitted from
the indicator.
Command R The indicator receives and sends weight data once time (the format is the same as
the continuous mode)
Command T The indicator receive the command and tare (the same as tare key); if no receipt of
the command. The indicator returns CR LF
Command Z The indicator receives the command and zero (the same as zero key); if no receipt
of the command, the indicator returns CR LF.
A12-51007 V1.0
-8-
Chapter 4 Calibration
4.1
Connect load cell properly, then turn on the indicator, press [#] key while it is initialization, it will
enter into the calibration mode and calibrate as following:
STEP
OPERATION
DISPLAY
for confirm
Example: 20
Select decimal point optional: 0~3, press [#] for
confirm
[P
NOTES
Example:3
Press [TARE] for selection of the digit bit; Press
[ZERO] for selection of the digit;
Press [#] for confirm the input of full range
Assure there is no load
End]
Press
the
calibration
switch under the lead
sealing board at the back
of the indicator
4.2 FAST CALIBRATION FOR ZERO POINT AND FULL RANGE POINT
Press [#] while it is initialization, it enters into the calibration mode.
4.2.1 Fast calibration for zero point:
At any time before it shows [nOLOAD], press [FUNC], it keeps the original division, decimal point, full
range and enter into the zero point calibration mode. Press [ZERO] when the stable signal is on, it
displays [End] and keeps the original full range point calibration. Press the calibration switch under
the lead sealing board at the back of the indicator, it saves the setting and back to the weighing status.
A12-51007 V1.0
-9-
EER 2
EER 3
EER 4
EER 5
The imputed weight is zero when full scale calibrated in calibrating mode.
EER 6
The unit weight is less than 0.25e when sampling in counting mode
bAt-lo
Low power
A12-51007 V1.0
- 10 -
Chapter 7 Maintenance
7.1 To guarantee indicator clarity and using life, the indicator shouldnt be placed directly under
sunshine and should be set in the plain space.
7.2 The indicator cant be placed into the place where the dust pollution and vibration are serious.
7.3 Load cell should connect with indicator reliably, and the system should be well connected into
ground. The indicator must be protected from high electrical field and high magnetic field.
In order to protect the operator, indicator and relevant device, you should mount lightning rod in
thunderstorm frequently happening area.
Load cell and indicator are static sensitive device, you must adopt anti static measures.
7.4 It is strictly forbidden to clean the case of indicator with intensive solvents (for example: benzene
and nitro oils)
7.5 Liquid and conducting particle should not be poured into the indicator, otherwise the electronic
components will be damaged and electric shock is likely to happen.
7.6 You should cut off power supply of indicator and relevant device before you pull-in and out the
connecting line of indicator and external device.
You must cut off power supply of indicator, before pulling out connecting line of load cell.
7.7 During operation, if trouble occurs, operator must pull off the power supply plug immediately, and
user should return this indicator to our company for repair. Non-weighing manufacturer should not
repair it, or by yourself, otherwise further destruction may happen.
7.8 The storage is not granted the free repair guarantee, because it is easily exhausted products.
In order to prolong using life, please charge the cell fully before using it. If you dont use the
indicator for a long time, you must charge the cell every two month and for eight hours/each
charging time.
Moving or installation must be carefully taken and must avoid strong vibration, impact and bump
in order to protect the storage cell from being damaged.
7.9 From invoice date, the indicator has a one-year free repair period. If any non-artificially obstacle
about the indicator happens under correct using conditions within this period, the user is allowed
to send the product with its guarantee card (of the correct number) back to our corporation for free
repair. The indicator shouldnt be taken apart, otherwise free guarantee will be cancelled.
A12-51007 V1.0
- 11 -