Location via proxy:   [ UP ]  
[Report a bug]   [Manage cookies]                

Puritan Bennett 980

Download as pdf or txt
Download as pdf or txt
You are on page 1of 506

Operator's Manual

Puritan Bennett
980 Series Ventilator

TM

Copyright Information
COVIDIEN, COVIDIEN with logo, and positive results for life are U.S and internationally registered trademarks of COVIDIEN AG. All other brands are trademarks of a Covidien company or
of their respective owners.
2014 COVIDIEN.
The information contained in this manual is the sole property of Covidien and may not be
duplicated without permission. This manual may be revised or replaced by Covidien at any time
and without notice. Ensure this manual is the most current applicable version. If in doubt,
contact Covidiens technical support department or visit the Puritan Bennett product manual
web page at:
http://www.covidien.com
Click: Our Products > Respiratory & Monitoring Products > Brands > Puritan Bennett. Click
Sales and Support Center > Product Manuals and Directions for Use, then follow the prompts
to select the desired manual.
While the information set forth herein is believed to be accurate, it is not a substitute for the
exercise of professional judgement.
The ventilator should be operated and serviced only by trained professionals.
Covidiens sole responsibility with respect to the ventilator and software, and its use, is as
stated in the limited warranty provided.
Nothing in this document shall limit or restrict in any way Covidiens right to revise or otherwise
change or modify the equipment (including its software) described herein, without notice. In
the absence of an express, written agreement to the contrary, Covidien has no obligation to
furnish any such revisions, changes, or modifications to the owner or user of the equipment
(including its software) described herein.

Table of Contents
1

Introduction

1.1

Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-1
1.1.1 Related Documents . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-1
1.2
Global Symbol Definitions . . . . . . . . . . . . . . . . . . . . . . . . .1-2
1.3
Safety Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-3
1.3.1 Safety Symbol Definitions . . . . . . . . . . . . . . . . . . . . . . . . . . .1-3
1.3.2 Warnings Regarding Fire Hazards
. . . . . . . . . . . . . . . . . . . . .1-3
1.3.3 General Warnings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-4
1.3.4 Warnings Regarding Environment of Use . . . . . . . . . . . . . . . .1-7
1.3.5 Warnings Before Using Equipment
. . . . . . . . . . . . . . . . . . . .1-8
1.3.6 Warnings Regarding Electrical Power . . . . . . . . . . . . . . . . . . .1-8
1.3.7 Warnings Regarding Ventilator Settings . . . . . . . . . . . . . . . . .1-9
1.3.8 Warnings Regarding Hoses, Tubing, and Accessories . . . . . . .1-9
1.3.9 Warnings Regarding Gas Sources
. . . . . . . . . . . . . . . . . . . .1-11
1.3.10 Warnings Regarding Infection Control . . . . . . . . . . . . . . . . .1-12
1.3.11 Warnings Regarding Ventilator Maintenance . . . . . . . . . . . .1-13
1.3.12 Cautions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-13
1.3.13 Notes
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-14
1.4
Obtaining Technical Assistance . . . . . . . . . . . . . . . . . . . .1-15
1.4.1 Technical Services . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-15
1.4.2 On-Screen Help . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-18
1.5
Warranty Information . . . . . . . . . . . . . . . . . . . . . . . . . . .1-19
1.6
Manufacture Date
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-19
1.7
Manufacturer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-20
1.8
Electromagnetic Compatibility . . . . . . . . . . . . . . . . . . . .1-20
2

Product Overview

2.1
2.2
2.3
2.4
2.5
2.6

Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-1
Ventilator Description . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-2
Indications For Use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-4
Contraindications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-4
Components List . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-5
Product Views . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-6
GUI Front View . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-6
GUI Rear View . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-7
BDU Front View . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-8
BDU Rear View . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-10

2.6.1
2.6.2
2.6.3
2.6.4

2.6.5 Ventilator Side Views . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-14


Mounting Configurations . . . . . . . . . . . . . . . . . . . . . . . .2-15
Battery Backup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-15
Graphical User Interface . . . . . . . . . . . . . . . . . . . . . . . . . .2-16
2.9.1 Primary Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-16
2.10
GUI Controls and Indicators . . . . . . . . . . . . . . . . . . . . . . .2-16
2.10.1 Control Keys . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-16
2.10.2 Visual Indicators
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-17
2.10.3 On-screen Symbols and Abbreviations . . . . . . . . . . . . . . . . .2-20
2.10.4 Audible Indicators . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-23
2.11
Breath Delivery Unit . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-25
2.11.1 BDU Controls and Indicators . . . . . . . . . . . . . . . . . . . . . . . .2-25
2.11.2 Connectors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-35
2.12
Special Features . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-35
2.13
Color Definitions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-36
2.14
Pneumatic Diagrams . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-36
2.7
2.8
2.9

Installation

3.1
3.2
3.3

Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-1
Safety Reminders . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-1
Product Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-2
How to Assemble Ventilator Components . . . . . . . . . . . . . . .3-2
Product Power Sources . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-2
Product Placement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-4
Product Connectivity . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-5
Connecting the Ventilator to AC Power . . . . . . . . . . . . . . . .3-5
Connecting the Gas Supplies . . . . . . . . . . . . . . . . . . . . . . . . .3-7
Filter Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-10
Connecting the Patient Circuit . . . . . . . . . . . . . . . . . . . . . . .3-14
How to Install Accessories . . . . . . . . . . . . . . . . . . . . . . . .3-18
Batteries . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-18
Battery Testing
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-23
Battery Performance Test Results . . . . . . . . . . . . . . . . . . . . .3-23
Battery Life . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-25
Battery Disposal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-25
Flex Arm . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-25
Humidifier . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-27
Ventilator Operating Modes . . . . . . . . . . . . . . . . . . . . . .3-30
Normal Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-30
Quick Start . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-30

3.3.1
3.3.2
3.4
3.5
3.5.1
3.5.2
3.5.3
3.5.4
3.6
3.6.1
3.6.2
3.6.3
3.6.4
3.6.5
3.6.6
3.6.7
3.7
3.7.1
3.7.2

ii

3.7.3 Stand-By State . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-31


3.7.4 Service Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-33
3.8
Product Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . .3-35
3.8.1 Preparing the Ventilator for Use . . . . . . . . . . . . . . . . . . . . . .3-37
3.8.2 Configuring the GUI . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-37
3.9
Installation Testing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-46
3.9.1 SST (Short Self Test) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-46
3.9.2 EST (Extended Self Test) . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-52
3.9.3 EST Test Sequence . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-54
3.9.4 EST Test Results . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-57
3.10
Operation Verification . . . . . . . . . . . . . . . . . . . . . . . . . . .3-58
4

Operation

4.1
4.2
4.3
4.4
4.5

Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-1
Ventilator Function . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-1
Ventilator Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-2
User Interface Management . . . . . . . . . . . . . . . . . . . . . . .4-2
Ventilator Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-7
Ventilator Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-10
Apnea Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-15
Alarm Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-16
Alarm Screen During Operation . . . . . . . . . . . . . . . . . . . . . .4-18
Making Ventilator Settings Changes . . . . . . . . . . . . . . . . . .4-19
Constant Timing Variable During Respiratory Rate Changes .4-20
Predicted Body Weight (PBW) Calculation . . . . . . . . . . .4-21
Non-invasive Ventilation (NIV) . . . . . . . . . . . . . . . . . . . .4-22
NIV Intended Use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-22
NIV Breathing Interfaces . . . . . . . . . . . . . . . . . . . . . . . . . . .4-22
NIV Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-23
Conversion from INVASIVE to NIV Vent Type . . . . . . . . . . . .4-24
Conversion from NIV to INVASIVE Vent Type . . . . . . . . . . . .4-25
High Spontaneous Inspiratory Time Limit Setting . . . . . . . . .4-26
NIV Apnea Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-27
NIV Alarm Settings
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-27
Manual Inspiration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-28
Respiratory Mechanics Maneuvers . . . . . . . . . . . . . . . . .4-28
Inspiratory Pause Maneuver . . . . . . . . . . . . . . . . . . . . . . . . .4-30
Expiratory Pause Maneuver . . . . . . . . . . . . . . . . . . . . . . . . .4-31
Other Respiratory Maneuvers . . . . . . . . . . . . . . . . . . . . . . . .4-32
Oxygen Sensor Function . . . . . . . . . . . . . . . . . . . . . . . . .4-32

4.5.1
4.5.2
4.5.3
4.5.4
4.5.5
4.5.6
4.6
4.7
4.7.1
4.7.2
4.7.3
4.7.4
4.7.5
4.7.6
4.7.7
4.7.8
4.8
4.9
4.9.1
4.9.2
4.9.3
4.10

iii

4.10.1 Oxygen Sensor Life . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-33


4.10.2 Oxygen Sensor Calibration . . . . . . . . . . . . . . . . . . . . . . . . . .4-34
4.10.3 Oxygen sensor calibration testing
. . . . . . . . . . . . . . . . . . . .4-34
4.11
Ventilator Protection Strategies . . . . . . . . . . . . . . . . . . .4-35
4.11.1 Power on Self Test (POST) . . . . . . . . . . . . . . . . . . . . . . . . . .4-35
4.11.2 Technical Fault . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-35
4.11.3 SST . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-35
4.11.4 Procedure Error . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-36
4.11.5 Ventilation Assurance . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-36
4.11.6 Safety Valve Open (SVO) . . . . . . . . . . . . . . . . . . . . . . . . . . .4-36
4.11.7 Ventilator Inoperative (Vent Inop) . . . . . . . . . . . . . . . . . . . .4-37
4.12
Ventilator Shutdown . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-37
5

Product Data Output

5.1
5.2
5.3
5.4

5.6
5.7
5.8
5.9

Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-1
Language . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-1
Data Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-1
Data Transfer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-1
GUI Screen Capture . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-2
Communication Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-3
Comm Port Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . .5-4
Serial Commands
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-5
RSET Command . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-5
SNDA Command . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-6
SNDF Command . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-10
Communication Ports . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-17
Port Use
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-19
Retrieving Stored Data . . . . . . . . . . . . . . . . . . . . . . . . . . .5-20
Display Configurability
. . . . . . . . . . . . . . . . . . . . . . . . . .5-20
Printing Data or Screen Captures . . . . . . . . . . . . . . . . . .5-21
Connectivity to External Systems . . . . . . . . . . . . . . . . . .5-21

Performance

5.4.1
5.4.2
5.4.3
5.4.4
5.4.5
5.4.6
5.4.7
5.5
5.5.1

6.1
6.2
6.3
6.4

iv

Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-1
System Options . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-1
Environmental Considerations . . . . . . . . . . . . . . . . . . . . .6-1
Ventilator Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-1
6.4.1 Ventilation Type
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-2
6.4.2 Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-2

6.4.3 Breath Type . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-2


Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-4
6.5.1 Alarm Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-4
6.5.2 Alarm Reset Key . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-8
6.5.3 Alarm Silence Key . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-8
6.5.4 Alarm Volume Key . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-8
6.5.5 Alarm Testing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-9
6.5.6 Viewing Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-14
6.5.7 Alarm Delay . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-15
6.5.8 Alarm Handling . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-15
6.5.9 AC POWER LOSS Alarm . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-35
6.5.10 Apnea Alarm . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-35
6.5.11 CIRCUIT DISCONNECT Alarm . . . . . . . . . . . . . . . . . . . . . . . .6-35
6.5.12 Loss of Power Alarm . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-36
6.5.13 DEVICE ALERT Alarm . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-36
6.5.14 High Circuit Pressure (1PPEAK) Alarm . . . . . . . . . . . . . . . . . .6-36
6.5.15 High Delivered O2% (1O2%) Alarm . . . . . . . . . . . . . . . . . . .6-37
. . . . . . . . . . .6-38
6.5.16 High Exhaled Minute Volume (1VE TOT) Alarm
6.5.17 High Exhaled Tidal Volume (1VTE) Alarm . . . . . . . . . . . . . . .6-38
6.5.18 High Inspired Tidal Volume (1VTI) Alarm . . . . . . . . . . . . . . . .6-38
6.5.19 High Respiratory Rate 1(fTOT) Alarm . . . . . . . . . . . . . . . . . . .6-39
6.5.20 INSPIRATION TOO LONG Alarm . . . . . . . . . . . . . . . . . . . . . .6-39
6.5.21 Low Circuit Pressure (3PPEAK) Alarm . . . . . . . . . . . . . . . . . . .6-39
6.5.22 Low Delivered O2% (3O2%) Alarm . . . . . . . . . . . . . . . . . . .6-40
6.5.23 Low Exhaled Mandatory Tidal Volume (3VTE MAND) Alarm . .6-41
6.5.24 Low Exhaled Spontaneous Tidal Volume (3VTE SPONT) Alarm .6-41
. . . . . . .6-41
6.5.25 Low Exhaled Total Minute Volume (3VE TOT) Alarm
6.5.26 PROCEDURE ERROR Alarm
. . . . . . . . . . . . . . . . . . . . . . . . .6-42
6.5.27 Severe Occlusion Alarm . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-42
6.6
Monitored Patient Data . . . . . . . . . . . . . . . . . . . . . . . . . .6-42
6.6.1 Total Exhaled Minute Volume (VE TOT) . . . . . . . . . . . . . . . . .6-43
6.6.2 Exhaled Spontaneous Minute Volume (VE SPONT) . . . . . . . . .6-43
6.6.3 Exhaled Tidal Volume (VTE) . . . . . . . . . . . . . . . . . . . . . . . . .6-43
6.6.4 Proximal Exhaled Minute Volume (VE TOTY) . . . . . . . . . . . . . .6-44
6.6.5 Proximal Exhaled Tidal Volume (VTEY) . . . . . . . . . . . . . . . . . .6-44
6.6.6 Exhaled Spontaneous Tidal Volume (VTE SPONT) . . . . . . . . . .6-44
6.6.7 Exhaled Mandatory Tidal Volume (VTE MAND) . . . . . . . . . . . .6-44
6.6.8 Exhaled mL/kg Volume . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-44
6.6.9 Inspired Tidal Volume (VTI) . . . . . . . . . . . . . . . . . . . . . . . . . .6-44
6.6.10 Proximal Inspired Tidal Volume (VTIY) . . . . . . . . . . . . . . . . . .6-45
6.6.11 Delivered mL/kg Volume . . . . . . . . . . . . . . . . . . . . . . . . . . .6-45
6.5

6.6.12 I:E Ratio . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-45


6.6.13 Mean Circuit Pressure (PMEAN) . . . . . . . . . . . . . . . . . . . . . . .6-45
6.6.14 Peak Circuit Pressure (PPEAK) . . . . . . . . . . . . . . . . . . . . . . . .6-45
6.6.15 End Inspiratory Pressure (PI END) . . . . . . . . . . . . . . . . . . . . . .6-45
6.6.16 End Expiratory Pressure (PEEP) . . . . . . . . . . . . . . . . . . . . . . .6-46
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-46
6.6.17 Intrinsic PEEP (PEEPI)
6.6.18 PAV-based Intrinsic PEEP (PEEPI PAV) . . . . . . . . . . . . . . . . . . .6-46
6.6.19 Total PEEP (PEEPTOT) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-46
6.6.20 Plateau Pressure (PPL) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-46
6.6.21 Total Respiratory Rate (fTOT) . . . . . . . . . . . . . . . . . . . . . . . . .6-46
6.6.22 PAV-based Lung Compliance (CPAV) . . . . . . . . . . . . . . . . . .6-46
6.6.23 PAV-based Patient Resistance (RPAV) . . . . . . . . . . . . . . . . . .6-47
6.6.24 PAV-based Lung Elastance (EPAV) . . . . . . . . . . . . . . . . . . . . .6-47
6.6.25 Spontaneous Rapid Shallow Breathing Index (f/ VT) . . . . . . .6-47
6.6.26 Spontaneous Inspiratory Time Ratio (TI/TTOT) . . . . . . . . . . . .6-47
6.6.27 Spontaneous Inspiratory Time (TI SPONT) . . . . . . . . . . . . . . . .6-47
6.6.28 PAV-based Total Airway Resistance (RTOT) . . . . . . . . . . . . . .6-47
6.6.29 Static Compliance (CSTAT) and Static Resistance (RSTAT) . . . .6-48
6.6.30 Dynamic Compliance (CDYN) . . . . . . . . . . . . . . . . . . . . . . . .6-49
6.6.31 Dynamic Resistance (RDYN) . . . . . . . . . . . . . . . . . . . . . . . . . .6-50
6.6.32 C20/C . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-50
6.6.33 End Expiratory Flow (EEF) . . . . . . . . . . . . . . . . . . . . . . . . . . .6-50
6.6.34 Peak Spontaneous Flow (PSF) . . . . . . . . . . . . . . . . . . . . . . .6-50
6.6.35 Displayed O2% . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-50
7

Preventive Maintenance

7.1
7.2
7.3
7.4
7.5

Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7-1
Ventilator Operational Time . . . . . . . . . . . . . . . . . . . . . . .7-1
Preventive Maintenance Intervals . . . . . . . . . . . . . . . . . . .7-1
Surface Cleaning of Exterior Surfaces . . . . . . . . . . . . . . . .7-4
Component Cleaning and Disinfection . . . . . . . . . . . . . . .7-6
Condensate Vial Cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . .7-8
Expiratory Filter Cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . .7-9
Exhalation Valve Flow Sensor Assembly Disinfection
. . . . . .7-12
Exhalation Valve Flow Sensor Assembly Reassembly . . . . . . .7-20
Exhalation Valve Flow Sensor Assembly Replacement . . . . . .7-22
Storage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7-23
Component Sterilization . . . . . . . . . . . . . . . . . . . . . . . . .7-23
Service Personnel Preventive Maintenance . . . . . . . . . .7-25
Safety Checks . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7-25

7.5.1
7.5.2
7.5.3
7.5.4
7.5.5
7.5.6
7.6
7.7
7.8

vi

7.9
7.10
7.11

Inspection and Calibration . . . . . . . . . . . . . . . . . . . . . . . .7-25


Documentation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7-25
Storage for Extended Periods . . . . . . . . . . . . . . . . . . . . .7-26

Troubleshooting

8.1
8.2
8.3
8.4
8.5
8.6

Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8-1
Problem Categories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8-1
How to Obtain Ventilator Service . . . . . . . . . . . . . . . . . . .8-1
Used Part Disposal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8-1
Ventilator Logs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8-2
Diagnostic Codes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8-4

Accessories

9.1
9.2

Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9-1
General Accessory Information . . . . . . . . . . . . . . . . . . . . .9-2

10

Theory of Operations

10.1
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .10-1
10.2
Theoretical Principles . . . . . . . . . . . . . . . . . . . . . . . . . . . .10-3
10.3
Applicable Technology . . . . . . . . . . . . . . . . . . . . . . . . . . .10-3
10.4
Inspiration Detection and initiation . . . . . . . . . . . . . .10-4
10.4.1 Pressure Triggering . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .10-5
10.4.2 Flow Triggering . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .10-6
10.4.3 Time Triggers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .10-8
10.4.4 Operator-initiated Triggers . . . . . . . . . . . . . . . . . . . . . . . . . .10-8
10.4.5 IE Sync Trigger . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .10-8
10.5
Exhalation Detection and Initiation . . . . . . . . . . . . . .10-9
10.5.1 Airway Pressure Method . . . . . . . . . . . . . . . . . . . . . . . . . . .10-9
10.5.2 Percent Peak Flow Method . . . . . . . . . . . . . . . . . . . . . . . .10-10
10.5.3 Time-cycling Method . . . . . . . . . . . . . . . . . . . . . . . . . . . . .10-11
10.5.4 IE Sync Cycling
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .10-11
10.5.5 Backup Methods . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .10-12
10.5.6 Time Limit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .10-12
10.5.7 High Circuit Pressure Limit . . . . . . . . . . . . . . . . . . . . . . . . .10-12
10.5.8 High Ventilator Pressure Limit . . . . . . . . . . . . . . . . . . . . . .10-12
10.5.9 High Inspired Tidal Volume Limit . . . . . . . . . . . . . . . . . . . .10-12
10.6
Compliance and BTPS Compensation . . . . . . . . . . . . . .10-13
10.6.1 Compliance Compensation in Volume-based Breaths
. . . .10-13

vii

10.6.2 BTPS Compensation in Volume-based Breaths . . . . . . . . . .10-17


10.7
Mandatory Breath Delivery . . . . . . . . . . . . . . . . . . . . . .10-17
10.7.1 Volume Control (VC) . . . . . . . . . . . . . . . . . . . . . . . . . . . . .10-18
10.7.2 Pressure Control (PC) . . . . . . . . . . . . . . . . . . . . . . . . . . . . .10-19
10.7.3 VC+
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .10-20
10.7.4 Rise time % . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .10-21
10.7.5 Manual Inspiration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .10-22
10.8
Spontaneous Breath Delivery . . . . . . . . . . . . . . . . . . . .10-22
10.8.1 Pressure Support (PS) . . . . . . . . . . . . . . . . . . . . . . . . . . . . .10-25
10.8.2 Volume Support (VS) . . . . . . . . . . . . . . . . . . . . . . . . . . . . .10-25
10.8.3 Tube Compensation . . . . . . . . . . . . . . . . . . . . . . . . . . . . .10-27
10.8.4 Proportional Assist Ventilation (PAV+) . . . . . . . . . . . . . .10-31
10.9
A/C Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .10-31
10.9.1 Changing to A/C Mode . . . . . . . . . . . . . . . . . . . . . . . . . . .10-34
10.10
SIMV Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .10-34
10.10.1Changing to SIMV Mode . . . . . . . . . . . . . . . . . . . . . . . . .10-37
10.11
Spontaneous (SPONT) Mode . . . . . . . . . . . . . . . . . . . . .10-39
10.11.1Changing to SPONT Mode . . . . . . . . . . . . . . . . . . . . . . . .10-40
10.12
Apnea Ventilation
. . . . . . . . . . . . . . . . . . . . . . . . . . . . .10-41
10.12.1Apnea Detection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .10-41
10.12.2Transition to Apnea Ventilation . . . . . . . . . . . . . . . . . . . .10-43
10.12.3Settings Changes During Apnea Ventilation . . . . . . . . . . .10-43
10.12.4Resetting Apnea Ventilation . . . . . . . . . . . . . . . . . . . . . . .10-44
10.12.5Apnea Ventilation in SIMV . . . . . . . . . . . . . . . . . . . . . . . .10-44
10.12.6Phasing in New Apnea Intervals . . . . . . . . . . . . . . . . . . . .10-45
10.13
Detecting Occlusion and Disconnect . . . . . . . . . . . . . . .10-45
10.13.1Occlusion . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .10-45
10.13.2Disconnect . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .10-47
10.13.3Annunciating Occlusion and Disconnect Alarms . . . . . . . .10-49
10.14
Respiratory Mechanics . . . . . . . . . . . . . . . . . . . . . . . . . .10-49
10.14.1Inspiratory Pause . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .10-50
10.14.2Expiratory Pause . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .10-53
10.14.3Negative Inspiratory Force (NIF) Maneuver . . . . . . . . . . . .10-54
. . . . . . . . . . . . . . . . .10-55
10.14.4P0.1 Maneuver (Occlusion Pressure)
10.14.5Vital Capacity (VC) Maneuver . . . . . . . . . . . . . . . . . . . . . .10-56
10.15
Ventilator Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . .10-57
10.15.1Apnea Ventilation
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . .10-57
10.15.2Circuit Type and Predicted Body Weight (PBW) . . . . . . . . .10-58
10.15.3Vent Type . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .10-59
10.15.4Mode and Breath Type . . . . . . . . . . . . . . . . . . . . . . . . . . .10-60
10.15.5Respiratory Rate (f) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .10-61

viii

10.15.6Tidal Volume (VT) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .10-62


. . . . . . . . . . . . . . . . . . . . . .10-62
10.15.7Peak Inspiratory Flow (VMAX)
10.15.8Plateau Time (TPL) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .10-63
10.15.9Flow Pattern . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .10-63
10.15.10Flow Sensitivity (VSENS) . . . . . . . . . . . . . . . . . . . . . . . . . .10-63
. . . . . . . . . . . . . . . . . . . . . . .10-64
10.15.11Pressure Sensitivity (PSENS)
10.15.12Inspiratory Pressure (PI) . . . . . . . . . . . . . . . . . . . . . . . . . .10-64
10.15.13Inspiratory Time (TI) . . . . . . . . . . . . . . . . . . . . . . . . . . . .10-65
10.15.14Expiratory Time (TE) . . . . . . . . . . . . . . . . . . . . . . . . . . . .10-66
10.15.15I:E Ratio . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .10-66
10.15.16High Pressure (PH) in BiLevel . . . . . . . . . . . . . . . . . . . . . .10-66
10.15.17Low Pressure (PL) in BiLevel . . . . . . . . . . . . . . . . . . . . . . .10-66
. . . . . . . . . . . . . . . . . . . . . . . .10-66
10.15.18High Time (TH) in BiLevel
10.15.19Low Time (TL) in BiLevel . . . . . . . . . . . . . . . . . . . . . . . . .10-67
10.15.20TH:TL Ratio in BiLevel . . . . . . . . . . . . . . . . . . . . . . . . . . .10-67
10.15.21PEEP . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .10-67
10.15.22Pressure Support (PSUPP) . . . . . . . . . . . . . . . . . . . . . . . . .10-68
10.15.23Volume Support (VT SUPP) . . . . . . . . . . . . . . . . . . . . . . . .10-68
10.15.24% Supp in TC . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .10-68
10.15.25% Supp in PAV+ . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .10-68
10.15.26Rise Time % . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .10-69
10.15.27Expiratory Sensitivity (ESENS) . . . . . . . . . . . . . . . . . . . . . .10-69
10.15.28Disconnect Sensitivity (DSENS) . . . . . . . . . . . . . . . . . . . . .10-70
10.15.29High Spontaneous Inspiratory Time Limit (2TI SPONT) . . . .10-70
10.15.30Humidification Type . . . . . . . . . . . . . . . . . . . . . . . . . . . .10-71
10.15.31Humidifier Volume . . . . . . . . . . . . . . . . . . . . . . . . . . . . .10-71
10.16
Safety Net . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .10-71
10.16.1User Error . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .10-72
10.16.2Patient Related Problems . . . . . . . . . . . . . . . . . . . . . . . . .10-73
10.16.3System Related Problems . . . . . . . . . . . . . . . . . . . . . . . . .10-73
10.16.4Background Diagnostic System . . . . . . . . . . . . . . . . . . . . .10-74
10.17
Power On Self Test (POST) . . . . . . . . . . . . . . . . . . . . . . .10-76
10.18
Short Self Test (SST) . . . . . . . . . . . . . . . . . . . . . . . . . . . .10-77
10.19
Extended Self Test (EST)
. . . . . . . . . . . . . . . . . . . . . . . .10-77
11

Specifications

11.1
11.2
11.3
11.4

Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .11-1
Measurement Uncertainty . . . . . . . . . . . . . . . . . . . . . . . .11-1
Physical Characteristics
. . . . . . . . . . . . . . . . . . . . . . . . . .11-2
Electrical Specifications . . . . . . . . . . . . . . . . . . . . . . . . . .11-6

ix

11.5
Interface Requirements . . . . . . . . . . . . . . . . . . . . . . . . . .11-6
11.6
Environmental Specifications . . . . . . . . . . . . . . . . . . . . .11-7
11.7
Performance Specifications . . . . . . . . . . . . . . . . . . . . . . .11-8
11.7.1 Ranges and Resolutions . . . . . . . . . . . . . . . . . . . . . . . . . . . .11-8
11.8
Regulatory Compliance . . . . . . . . . . . . . . . . . . . . . . . . .11-28
11.9
Manufacturers Declaration . . . . . . . . . . . . . . . . . . . . . .11-30
11.10
Safety Tests . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .11-35
11.11
Essential Performance Requirements . . . . . . . . . . . . . .11-35
A

BiLevel 2.0 Appendix

A.1
A.2
A.3
A.4
A.5
A.6
A.7
A.8
A.9
A.9.1
A.9.2
A.9.3
A.9.4
A.10

Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-1
Intended Use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-2
Safety Symbol Definitions . . . . . . . . . . . . . . . . . . . . . . . . A-2
Setting Up BiLevel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-3
Using Pressure Support with BiLevel . . . . . . . . . . . . . . . A-4
Manual Inspirations in BiLevel Mode . . . . . . . . . . . . . . . A-5
Respiratory Mechanics Maneuvers in BiLevel . . . . . . . . A-6
Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-6
Technical Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-6
Synchrony in BiLevel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-7
Patient Monitoring in BiLevel . . . . . . . . . . . . . . . . . . . . . . . . A-8
APRV Strategy in BiLevel . . . . . . . . . . . . . . . . . . . . . . . . . . . A-9
Technical Structure of BiLevel
. . . . . . . . . . . . . . . . . . . . . . A-10
Mode Changes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-10

Leak Sync Appendix

B.1
B.2
B.3
B.4
B.5
B.6

Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . B-1
Intended Use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . B-1
Safety Symbol Definitions . . . . . . . . . . . . . . . . . . . . . . . . B-2
Leak Sync . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . B-2
Setting Up Leak Sync . . . . . . . . . . . . . . . . . . . . . . . . . . . . B-3
When Leak Sync is Enabled . . . . . . . . . . . . . . . . . . . . . . . B-4
Adjusting Disconnect Sensitivity (DSENS) . . . . . . . . . . . . . . . . .B-5
Monitored Patient Data . . . . . . . . . . . . . . . . . . . . . . . . . . . . .B-6
Technical Discussion . . . . . . . . . . . . . . . . . . . . . . . . . . . . . B-8
Inspired Tidal Volume (VTL) Accuracy During Leak Sync . . . . .B-8
Exhaled Tidal Volume (VTE) Accuracy During Leak Sync . . . . .B-8
%LEAK Calculation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .B-9
Circuit Disconnect Alarm During Leak Sync . . . . . . . . . . . . . .B-9

B.6.1
B.6.2
B.7
B.7.1
B.7.2
B.7.3
B.7.4

IE Sync Appendix

C.1
C.2
C.3
C.4
C.5
C.6
C.7
C.8
C.9

Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Intended Use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Safety Symbol Definitions . . . . . . . . . . . . . . . . . . . . . . . .
IE Sync Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Setting Up IE Sync
..............................
Setting Up IE Sync for Monitoring Only . . . . . . . . . . . . .
IE Sync Monitoring Display . . . . . . . . . . . . . . . . . . . . . . .
Technical Description . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Alarms during IE Sync
...........................

PAV+ Appendix

D.1
D.2
D.3
D.4
D.4.1
D.4.2
D.4.3
D.4.4
D.4.5
D.4.6
D.4.7
D.4.8
D.5
D.5.1
D.5.2
D.5.3
D.5.4
D.5.5

Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . D-1
Intended Use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . D-1
Safety Symbol Definitions . . . . . . . . . . . . . . . . . . . . . . . . D-2
PAV+ . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . D-3
Setting Up PAV+ . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . D-5
PBW and Tube ID
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . D-6
Apnea Parameters Adjustment
. . . . . . . . . . . . . . . . . . . . . . D-7
Alarm Settings Adjustment . . . . . . . . . . . . . . . . . . . . . . . . . D-8
PAV+ Ventilator Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . D-8
PAV+ Alarm Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . D-8
Monitored Data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . D-9
PAV+ Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . D-10
Ventilator Settings/Guidance . . . . . . . . . . . . . . . . . . . . D-10
Specified performance . . . . . . . . . . . . . . . . . . . . . . . . . . . . D-11
Graphics Displays in PAV+ . . . . . . . . . . . . . . . . . . . . . . . . . D-11
WOB Terms and Definitions . . . . . . . . . . . . . . . . . . . . . . . . D-12
Technical Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . D-14
Protection Against Hazard . . . . . . . . . . . . . . . . . . . . . . . . . D-19

NeoMode 2.0 Appendix

E.1
E.2
E.3
E.4
E.5
E.6

Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .E-1
Intended Use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .E-1
Safety Symbol Definitions . . . . . . . . . . . . . . . . . . . . . . . . .E-1
Safety Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .E-2
Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .E-3
Neonatal Door and Filter Installation . . . . . . . . . . . . . . . .E-3

C-1
C-1
C-1
C-2
C-2
C-4
C-5
C-7
C-7

xi

E.7
E.8
E.9
E.9.1
E.9.2
E.9.3
E.9.4
E.9.5
E.9.6
E.9.7
F

Proximal Flow Option Appendix

F.1
F.2
F.3
F.4
F.5
F.6

Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .F-1
Intended Use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .F-1
Safety Symbol Definitions . . . . . . . . . . . . . . . . . . . . . . . . .F-1
Software/Hardware Requirements . . . . . . . . . . . . . . . . . .F-2
Safety Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .F-2
Proximal Flow Option Description . . . . . . . . . . . . . . . . . .F-5
Proximal Flow Option components
. . . . . . . . . . . . . . . . . . . . F-5
On-screen symbols . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .F-6
Sensor Calibration and Sensor Line Purging . . . . . . . . . .F-8
SST Requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .F-9
Attaching the Proximal Flow Sensor for SST . . . . . . . . . . . . . F-11
Disabling/Enabling the Proximal Flow Option . . . . . . . .F-12
Using the Proximal Flow Sensor . . . . . . . . . . . . . . . . . . .F-13
How to Perform a Manual Purge . . . . . . . . . . . . . . . . . . . . . F-15
Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .F-16
Ranges, Resolutions, and Accuracies
. . . . . . . . . . . . . . .F-17
Proximal Flow Sensor Specifications . . . . . . . . . . . . . . . . . . . F-17
Part Numbers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .F-18

F.6.1
F.7
F.8
F.9
F.9.1
F.10
F.11
F.11.1
F.12
F.13
F.13.1
F.14
Glossary

xii

How to Empty the Condensate Vial . . . . . . . . . . . . . . . . .E-5


Connecting the Breathing Circuit . . . . . . . . . . . . . . . . . . .E-6
Ventilation Features . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .E-8
Predicted Body Weight (PBW) vs. Patient Length . . . . . . . . . .E-8
Elevate O2 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .E-8
CPAP Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .E-9
Entering CPAP From Other Ventilation Modes . . . . . . . . . . .E-10
Exiting CPAP Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .E-11
Compliance Compensation . . . . . . . . . . . . . . . . . . . . . . . . .E-11
Settings, Alarms, and Monitored Patient Data . . . . . . . . . . .E-11

List of Tables
Table1-1.
Table1-2.
Table2-1.
Table2-2.
Table2-3.
Table2-4.
Table2-5.
Table2-6.
Table2-7.
Table2-8.
Table2-9.
Table2-10.
Table2-11.
Table3-1.
Table3-2.
Table3-3.
Table3-4.
Table3-5.
Table3-6.
Table3-7.
Table3-8.
Table3-9.
Table4-1.
Table4-2.
Table4-3.
Table4-4.
Table5-1.
Table5-2.
Table6-1.
Table6-2.
Table6-3.
Table6-4.
Table6-5.
Table6-6.
Table7-1.
Table7-2.
Table7-3.
Table7-4.
Table7-5.

Shipping Carton Symbols and Descriptions . . . . . . . . . . . . . . . . . . 1-2


Safety Symbol Definitions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-3
Typical Packing List . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-5
BDU Front Label Symbols and Descriptions . . . . . . . . . . . . . . . . . . 2-9
BDU Rear Label or Panel Symbols and Descriptions. . . . . . . . . . 2-11
Common Symbols Found on GUI or BDU Labels . . . . . . . . . . . . . 2-13
GUI Control Keys . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-16
GUI Visual Indicators . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-18
Symbols and Abbreviations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-21
GUI Audible Indicator Functions . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-24
Status Display Indicators and Descriptions . . . . . . . . . . . . . . . . . . 2-30
BDU Audible Indicator Functions . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-35
Color Legend . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-36
Patient Types and PBW Values . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-15
Ventilator Configuration. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-35
SST Test Sequence . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-49
Humidifier Volumes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-50
Individual SST Results . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-51
Overall SST Outcomes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-51
EST Test Sequence . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-55
Individual EST Results . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-57
Overall EST Outcomes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-58
Gestures and Their Meanings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-6
Setting Up a Patient for NIV. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-23
INVASIVE to NIV on Same Patient. . . . . . . . . . . . . . . . . . . . . . . . . . . 4-24
NIV to INVASIVE on Same Patient . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-25
MISCA Response . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-6
MISCF Response . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-11
Alarm Descriptions and Symbols. . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-6
Alarm Prioritization . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-16
Technical Alarm Categories. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-17
Technical Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-18
Non-technical Alarm Summary. . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-18
Non-Technical Alarms and Suggested Responses . . . . . . . . . . . 6-29
Operator Preventive Maintenance Frequency . . . . . . . . . . . . . . . . 7-2
Surface Cleaning Agents . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-5
Component Cleaning Agents and Disinfection Procedures . . . 7-6
Sterilization Parameters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-24
Component Sterilization Procedures. . . . . . . . . . . . . . . . . . . . . . . . 7-24

xiii

Table9-1.
Table10-1.
Table10-2.
Table10-3.
Table10-4.
Table10-5.
Table10-6.
Table10-7.
Table10-8.
Table10-9.
Table10-10.
Table11-1.
Table11-2.
Table11-3.
Table11-4.
Table11-5.
Table11-6.
Table11-7.
Table11-8.
Table11-9.
Table11-10.
Table11-11.
Table11-12.
Table11-13.
Table11-14.
Table11-15.
Table11-16.
Table11-17.
Table11-18.
TableA-1.
TableB-1.
TableB-2.
TableB-3.
TableC-1.
TableD-1.
TableD-2.
TableD-3.
TableD-4.
TableE-1.
TableE-2.
TableE-3.

xiv

Covidien Accessories and Options . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-4


Compliance Volume Factors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .10-16
Maximum Pressure Adjustments . . . . . . . . . . . . . . . . . . . . . . . . . .10-21
Spontaneous Breath Delivery Characteristics . . . . . . . . . . . . . . .10-23
Maximum Pressure Adjustments. . . . . . . . . . . . . . . . . . . . . . . . . . .10-26
Inspiratory and Expiratory Pause Events. . . . . . . . . . . . . . . . . . . .10-51
Values for VT Based on Circuit Type . . . . . . . . . . . . . . . . . . . . . . . .10-58
Peak Flow and Circuit Type (Leak Sync Disabled) . . . . . . . . . . .10-59
Modes and Breath Types . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .10-60
Safety PCV Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .10-72
Inspiratory Backup Ventilation Settings . . . . . . . . . . . . . . . . . . . .10-75
Performance Verification Equipment Uncertainty . . . . . . . . . . . 11-1
Physical Characteristics. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11-2
Pneumatic Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11-3
Technical Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11-3
Electrical Specifications. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11-6
Interface Pin Designations. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11-6
Nurse Call Pin Designations. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11-7
Environmental Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11-7
Ventilator Settings Range and Resolution . . . . . . . . . . . . . . . . . . . 11-8
Alarm Settings Range and Resolution . . . . . . . . . . . . . . . . . . . . .11-17
Patient Data Range and Resolution . . . . . . . . . . . . . . . . . . . . . . . .11-19
Delivery Accuracy . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .11-26
Monitoring (Patient Data) Accuracy . . . . . . . . . . . . . . . . . . . . . . . .11-27
Computed Value Accuracy . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .11-28
Electromagnetic Emissions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .11-30
Electromagnetic Immunity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .11-31
Recommended Separation Distances for RF . . . . . . . . . . . . . . .11-34
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .11-35
Safety Symbol Definitions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-3
Safety Symbol Definitions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . B-2
Leak Compensation Volumes Based on Patient Type . . . . . . . . . B-3
DSENS settings. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . B-6
Safety Symbol Definitions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . C-2
Safety Symbol Definitions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .D-2
Absolute limits for PAV+ Monitored Data . . . . . . . . . . . . . . . . . . . .D-9
PAV+ Work of Breathing terms. . . . . . . . . . . . . . . . . . . . . . . . . . . . . D-13
Default PBW-based Resistance Values. . . . . . . . . . . . . . . . . . . . . . D-23
Safety Symbol Definitions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . E-2
Recommended Breathing Circuits . . . . . . . . . . . . . . . . . . . . . . . . . . . E-8
Delivered Volume Accuracy . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . E-12

TableE-4.
Monitored Inspired Volume (VTI) Accuracy . . . . . . . . . . . . . . . . . . E-12
Monitored Exhaled Tidal Volume (VTE) Accuracy . . . . . . . . . . . . E-13
TableE-5.
TableF-1.
Safety Symbol Definitions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . F-2
TableF-2.
Proximal Flow Option Patient Data Symbols . . . . . . . . . . . . . . . . . F-8
TableF-3.
Proximal Flow Option SST test Sequence. . . . . . . . . . . . . . . . . . . . F-10
TableF-4.
Proximal Flow Sensor Volume Accuracy. . . . . . . . . . . . . . . . . . . . . F-17
TableF-5.
Proximal Flow Sensor Specifications . . . . . . . . . . . . . . . . . . . . . . . . F-18
TableF-6.
Proximal Flow Option Component Part Numbers. . . . . . . . . . . . F-18
TableGlossary-1.Glossary of Ventilation Terms . . . . . . . . . . . . . . . . . . . . . Glossary-1
TableGlossary-2.Units of Measure. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Glossary-9
TableGlossary-3.Technical Abbreviations . . . . . . . . . . . . . . . . . . . . . . . . .Glossary-10

xv

Page Left Intentionally Blank

xvi

List of Figures
Figure2-1.
Figure2-2.
Figure2-3.
Figure2-4.
Figure2-5.
Figure2-6.
Figure2-7.
Figure2-8.
Figure2-9.
Figure2-10.
Figure2-11.
Figure2-12.
Figure3-1.
Figure3-2.
Figure3-3.
Figure3-4.
Figure3-5.
Figure3-6.
Figure3-7.
Figure3-8.
Figure3-9.
Figure3-10.
Figure3-11.
Figure3-12.
Figure3-13.
Figure3-14.
Figure3-15.
Figure4-1.
Figure4-2.
Figure4-3.
Figure4-4.
Figure4-5.
Figure4-6.
Figure4-7.
Figure4-8.
Figure4-9.
Figure4-10.
Figure4-11.
Figure4-12.

GUI Front View ............................................................................................2-6


GUI Rear View ..............................................................................................2-7
BDU Front View ..........................................................................................2-8
BDU Rear View ......................................................................................... 2-10
Installed Software Options .................................................................. 2-11
Ventilator Right Side View ................................................................... 2-14
Ventilator Left Side View ...................................................................... 2-15
Ventilator Power Switch and AC Indicator .................................... 2-26
Service Mode Button (TEST) ................................................................ 2-27
Sample Status Display During Normal Ventilation ..................... 2-29
Pneumatic Diagram ............................................................................... 2-37
Pneumatic Diagram Optional Proximal Flow System .......... 2-38
Example of Freestanding Ventilator Placement .............................3-5
Power Cord Retainer on BDU ................................................................3-7
Connecting the Ventilator to the Gas Supplies ..............................3-9
Adult/Pediatric Filter Installation ...................................................... 3-12
Installing the Neonatal Filter .............................................................. 3-13
Drain Bag ................................................................................................... 3-14
Connecting the Adult or Pediatric Patient Circuit ...................... 3-16
Connecting the Neonatal Patient Circuit ....................................... 3-17
Ventilator Battery .................................................................................... 3-19
Proper Battery Orientation .................................................................. 3-20
Battery Compartment Locations ....................................................... 3-21
Flex Arm Installation .............................................................................. 3-26
Bracket Installation on Rail .................................................................. 3-28
Humidifier Installation to Ventilator ................................................ 3-29
Service Mode Button (TEST) ................................................................ 3-34
Areas of the GUI ..........................................................................................4-3
Pushpin Icon ................................................................................................4-5
New Patient Settings ................................................................................4-9
Open Menu Tab ....................................................................................... 4-11
New Patient setup Screen ................................................................... 4-12
Apnea Setup Screen .............................................................................. 4-15
Alarms Settings Screen ......................................................................... 4-17
Alarm Screen During Operation ........................................................ 4-19
Vent Setup Button NIV Indicating NIV vent type ..................... 4-25
2TI SPONT Indicator .................................................................................. 4-26
Default NIV Alarm Settings .................................................................. 4-27
RM in Menu Tab ....................................................................................... 4-29

xvii

Figure4-13. Respiratory Maneuver Tabs ................................................................ 4-29


Figure4-14. More Settings Screen with O2 Sensor Enabled ............................ 4-33
Figure5-1. Comm Setup Screen .................................................................................5-4
Figure5-2. Port Locations .......................................................................................... 5-18
Figure6-1. Alarm Message Format ............................................................................6-5
Figure7-1. Condensate Vial .........................................................................................7-9
Figure7-2. Filter Top and Bottom Views ............................................................. 7-10
Figure7-3. From Patient Port .................................................................................... 7-11
Figure7-4. Expiratory Filter Drying Orientation ................................................ 7-12
Figure7-5. Exhalation Valve Flow Sensor Assembly ........................................ 7-14
Figure7-6. Exhalation Valve Flow Sensor Components ................................. 7-14
Figure7-7. Exhalation Valve Flow Sensor Assembly Removal ...................... 7-15
Figure7-8. Exhalation Valve Diaphragm Removal ............................................ 7-15
Figure7-9. Expiratory Filter Seal Removal ........................................................... 7-16
Figure7-10. Pressure Sensor Filter Removal .......................................................... 7-16
Figure7-11. Immersion Method ................................................................................ 7-18
Figure7-12. Exhalation Valve Flow Sensor Assembly Reprocessing Kit ...... 7-20
Figure7-13. Installing the Pressure Sensor Filter ................................................. 7-20
Figure7-14. Installing the Expiratory Filter Seal ................................................... 7-21
Figure7-15. Installing the Diaphragm ..................................................................... 7-22
Figure7-16. Installing the Exhalation Valve Flow Sensor Assembly ............. 7-23
Figure8-1. Log Screen ....................................................................................................8-4
Figure9-1. Ventilator with Accessories ....................................................................9-3
Figure9-2. Additional Accessories .............................................................................9-4
Figure10-1. Inspiration Using Pressure Sensitivity ............................................. 10-6
Figure10-2. Inspiration Using Flow Sensitivity ..................................................... 10-7
Figure10-3. Breath Activity During Time-triggered Inspiration ..................... 10-8
Figure10-4. Exhalation via the Airway Pressure Method ............................... 10-10
Figure10-5. Exhalation via the Percent Peak Flow Method ........................... 10-11
Figure10-6. Square Flow Pattern ............................................................................ 10-13
Figure10-7. Descending Ramp Flow Pattern ...................................................... 10-14
Figure10-8. Ideal Waveform Using Square Flow Pattern ............................... 10-18
Figure10-9. Ideal Waveform Using Descending Ramp Flow Pattern ........ 10-19
Figure10-10. Ideal Waveform Using Pressure Control Ventilation ................ 10-20
Figure10-11. ET Tube Target Pressure vs. Flow .................................................... 10-30
Figure10-12. Tracheostomy Tube Target Pressure vs. Flow ............................ 10-31
Figure10-13. No Patient Inspiratory Effort Detected .......................................... 10-32
Figure10-14. Patient Inspiratory Effort Detected ................................................ 10-33
Figure10-15. Combined VIM and PIM Breaths ...................................................... 10-33
Figure10-16. Mandatory and Spontaneous Intervals ........................................ 10-34
Figure10-17. PIM Delivered Within Mandatory Interval ................................... 10-35

xviii

Figure10-18. PIM Not Delivered Within Mandatory Interval ........................... 10-35


Figure10-19. Apnea Interval Equals Breath Period ............................................. 10-42
Figure10-20. Apnea Interval Greater Than Breath Period ................................ 10-42
Figure10-21. Apnea Interval Less Than Breath Period ....................................... 10-43
Figure10-22. Apnea Ventilation in SIMV ................................................................. 10-45
FigureA-1. Spontaneous Breathing at PL ............................................................... A-1
FigureA-2. BiLevel Mode .............................................................................................. A-2
FigureA-3. BiLevel Setup Screen ............................................................................... A-4
FigureA-4. BiLevel with Pressure Support ............................................................. A-5
FigureA-5. Spontaneous and Synchronous Intervals ........................................ A-8
FigureA-6. APRV With Spontaneous Breathing at PH ........................................ A-9
FigureB-1. Enabling Leak Sync ...................................................................................B-4
FigureB-2. GUI Screen when Leak Sync is Enabled .............................................B-5
FigureB-3. Leak Sync Monitored Patient Data ......................................................B-7
FigureB-4. Circuit Disconnect During VC ................................................................B-9
FigureC-1. Setup Screen ................................................................................................C-3
FigureC-2. Monitoring Setup ......................................................................................C-4
FigureC-3. Monitoring Dialog .....................................................................................C-5
FigureC-4. Monitoring Waveform .............................................................................C-6
FigureD-1. Ventilator Setup Screen .......................................................................... D-6
FigureD-2. Graphics displays in PAV+ ...................................................................D-14
FigureD-3. Use of Default Lung Resistance ........................................................D-23
FigureE-1. Installing Neonatal Filter and Door ..................................................... E-4
FigureE-2. How to Connect the Breathing Circuit ............................................... E-7
FigureE-3. CPAP Setup Screen ..................................................................................E-10
FigureF-1. Proximal Flow Sensor ............................................................................... F-6
FigureF-2. Sample GUI screen Showing Proximal Flow Data .......................... F-7
FigureF-3. Message During Autozero and Purge Processes ............................ F-9
FigureF-4. Attaching Proximal Flow Sensor to Ventilator ..............................F-11
FigureF-5. Enabling/disabling Proximal Flow Sensor ......................................F-13
FigureF-6. Attaching Proximal Flow Sensor ........................................................F-14
FigureF-7. Manual Purge ............................................................................................F-16
FigureF-8. Alarm Message Prox Inoperative .................................................F-17

xix

Page Left Intentionally Blank

xx

1 Introduction

1.1

Overview
This manual contains information for operating the Puritan Bennett 980
Series Ventilators. Before operating the ventilator system, thoroughly read this
manual. The latest version of this manual is available on the Internet at:
http://www.covidien.com

To order an additional copy of this manual, contact Covidien Customer Service


or your local representative.
1.1.1

Related Documents
Documentation is available online at the URL above. Covidien makes available
all appropriate information relevant to use and service of the ventilator. For
further assistance, contact your local Covidien representative.

The Puritan Bennett 980 Series Ventilator Operators Manual Provides


basic information on operating the ventilator and troubleshooting errors or malfunctions. Before using the ventilator, thoroughly read this manual.

The Puritan Bennett 980 Series Ventilator Service Manual Provides information to Covidien-trained service technicians for use when testing, troubleshooting, repairing, and upgrading the ventilator.

This chapter contains the following:

Symbol definitions

Safety Information, including Warnings, Cautions, and Notes

Technical assistance information

How to access on-screen Help

How to access warranty information

Serial number interpretation

Information regarding Electromagnetic susceptibility

1-1

Introduction

1.2

Global Symbol Definitions


The following table describes the symbols shown on the ventilator shipping
cartons. Other symbols appearing on various labels are shown in Chapter 2.
Table1-1.Shipping Carton Symbols and Descriptions
Symbol

Description
Serial number
Part number

Manufacturer

This side up

Fragile

Humidity limitations: 10% to 95% relative humidity, non-condensing (operation and storage)

Temperature limitations: 10C to 40C (50F to 104F) (operation)


-20C to 70C (-68F to 158F) (storage)

Atmospheric pressure limitations: 70 kPa to 106 kPa (10.2 psi to


15.4 psi)

Keep dry

CSA certification mark that signifies the product has been evaluated
to the applicable ANSI/Underwriters Laboratories Inc. (UL) and CSA
standards for use in the US and Canada.

1-2

Operators Manual

Safety Information

Table1-1.Shipping Carton Symbols and Descriptions (Continued)


Symbol

Description
US federal law restricts this device to sale by or on the order of a
physician.

Refer to instruction manual.

1.3

1.3.1

Safety Information
Safety Symbol Definitions
This section contains safety information for users, who should always exercise
appropriate caution while using the ventilator.
Table1-2.Safety Symbol Definitions
Symbol

Definition
WARNING
Warnings alert users to potential serious outcomes (death, injury, or adverse
events) to the patient, user, or environment.
Caution
Cautions alert users to exercise appropriate care for safe and effective use of the
product.
Note
Notes provide additional guidelines or information.

1.3.2

Warnings Regarding Fire Hazards


WARNING:
Explosion hazard Do not use in the presence of flammable gases. An
oxygen-rich environment accelerates combustibility.

Operators Manual

1-3

Introduction

1.3.3

1-4

WARNING:
To avoid a fire hazard, keep all components of the system away from all
sources of ignition (such as matches, lighted cigarettes, flammable medical
gases, and/or heaters). Oxygen-rich environments accelerate combustibility.
WARNING:
In case of fire or a burning smell, immediately take the following actions if it
is safe to do so: disconnect the patient from the ventilator and disconnect the
ventilator from the oxygen supply, facility power, and all batteries. Provide
alternate method of ventilatory support to the patient, if required.
WARNING:
Replacement of batteries by inadequately trained personnel could result in an
unacceptable risk, such as excessive temperatures, fire, or explosion.
WARNING:
To minimize fire hazard, inspect and clean or replace, as necessary, any
damaged ventilator parts that come into contact with oxygen.
WARNING:
To prevent electrostatic discharge (ESD) and potential fire hazard, do not use
antistatic or electrically conductive hoses or tubing in or near the ventilator
breathing system.

General Warnings
WARNING:
To ensure proper operation and avoid the possibility of physical injury, only
qualified medical personnel should attempt to set up the ventilator and
administer treatment with the ventilator.
WARNING:
In case of ventilator failure, the lack of immediate access to appropriate
alternative means of ventilation can result in patient death. An alternative
source of ventilation, such as a self-inflating, manually-powered resuscitator
(as specified in ISO 10651-4 with mask) should always be available when
using the ventilator.

Operators Manual

Safety Information

WARNING:
Patients on mechanical ventilation should be monitored by clinicians for
proper patient ventilation.
WARNING:
The ventilator system is not intended to be a comprehensive monitoring
device and does not activate alarms for all types of conditions. For a detailed
understanding of ventilator operations, be sure to thoroughly read this
manual before attempting to use the ventilator system.
WARNING:
To prevent patient injury, do not use the ventilator if it has a known
malfunction. Never attempt to override serious malfunctions. Replace the
ventilator and have the faulty unit repaired by trained service personnel.
WARNING:
To prevent patient injury, do not make unauthorized modifications to the
ventilator.
WARNING:
To prevent injury and avoid interfering with ventilator operation, do not
insert tools or any other objects into any of the ventilators openings or ports.
WARNING:
The audio alarm volume level is adjustable. The operator should set the
volume at a level that allows the operator to distinguish the audio alarm
above background noise levels. Reference To adjust alarm volume, p. 3-41 for
instructions on alarm volume adjustment.
WARNING:
Do not silence, disable, or decrease the volume of the ventilators audible
alarm if patient safety could be compromised.
WARNING:
If increased pressures are observed during ventilation, it may indicate a
problem with the ventilator. Check for blocked airway, circuit occlusion,
and/or run SST.

Operators Manual

1-5

Introduction

1-6

WARNING:
The LCD panel contains toxic chemicals. Do not touch broken LCD panels.
Physical contact with a broken LCD panel can result in transmission or
ingestion of toxic substances.
WARNING:
If the Graphical User Interface (GUI) display/LCD panel is blank or experiences
interference and cannot be read, check the patient, then verify via the status
display that ventilation is continuing as set. Because breath delivery is
controlled independently from the GUI, problems with the display will not, by
themselves, affect ventilation. The ventilator, however, should be replaced as
soon as possible and repaired by qualified service personnel.
WARNING:
The Puritan Bennett 980 Series Ventilator contains phthalates. When used
as indicated, very limited exposure to trace amounts of phthalates may occur.
There is no clear clinical evidence that this degree of exposure increases
clinical risk. However, in order to minimize risk of phthalate exposure in
children and nursing or pregnant women, this product should only be used as
directed.
WARNING:
Even though the 980 Series Ventilator meets the standards listed in
Chapter 11, the internal Lithium-ion battery of the device is considered to
be Dangerous Goods (DG) Class 9 - Miscellaneous, when transported in
commerce. The 980 Series Ventilator and/or the associated Lithium-ion
battery are subject to strict transport conditions under the Dangerous
Goods Regulation for air transport (IATA: International Air Transport
Association), International Maritime Dangerous Goods code for sea and
the European Agreement concerning the International Carriage of
Dangerous Goods by Road (ADR) for Europe. Private individuals who
transport the device are excluded from these regulations although for air
transport some requirements may apply.

Operators Manual

Safety Information

1.3.4

Warnings Regarding Environment of Use


WARNING:
Do not position the ventilator next to anything that blocks or restricts the gas
inlet or cooling air circulation openings, gas exhaust port, fan intake, or alarm
speaker, as this may:

limit the air circulation around the ventilator, potentially causing overheating;

limit the ventilator's ability to exhaust patient exhaled gas leading to


potential harm;

limit the clinicians ability to hear ventilator alarms.

WARNING:
To avoid injury, do not position the ventilator in a way that makes it difficult
to disconnect the patient.
WARNING:
To ensure proper operation, do not position the ventilator in a way that
makes it difficult to access the AC power cord.
WARNING:
Do not use the ventilator in a hyperbaric chamber. It has not been validated
for use in this environment.
WARNING:
Do not use the ventilator in the presence of strong magnetic fields. Doing so
could cause a ventilator malfunction.
WARNING:
Do not use the ventilator during radiotherapy (i.e. cancer treatment using
ionizing radiation), as doing so could cause a ventilator malfunction.
WARNING:
To avoid the risk of ventilator malfunction, operate the ventilator in an
environment that meets specifications. Reference Environmental
Specifications, p. 11-7.

Operators Manual

1-7

Introduction

1.3.5

1.3.6

1-8

WARNING:
Do not use the ventilator as an EMS transport ventilator. It has not been
approved or validated for this use.

Warnings Before Using Equipment


WARNING:
Before activating any part of the ventilator, be sure to check the equipment
for proper operation and, if appropriate, run SST as described in this manual.
Reference To run SST, p. 3-48.
WARNING:
Check for leaks in the ventilator breathing system by running SST prior to
ventilating a patient.
WARNING:
Lock the ventilators casters during use to avoid the possibility of extubation
due to inadvertent ventilator movement.
WARNING:
The ventilator accuracies listed in the Ventilator Settings, Alarm Settings, and
Patient Data tables in Chapter 11 are applicable only under specified
operating conditions. Reference Environmental Specifications, p. 11-7. If the
ventilator is operated outside specified ranges, the ventilator may supply
incorrect information and the accuracies listed in the aforementioned tables
do not apply. A hospital Biomedical Technician must verify the ventilator is
operated in the environmental conditions specified.

Warnings Regarding Electrical Power


WARNING:
To avoid the risk of electrical shock:

Use only Covidien-branded batteries, adapters, and cables.

Do not use batteries, adapters or cables with visible signs of damage.

Do not touch internal components.

Operators Manual

Safety Information

1.3.7

1.3.8

Warnings Regarding Ventilator Settings


WARNING:
The ventilator offers a variety of breath delivery options. Throughout the
patient's treatment, the clinician should carefully select the ventilation mode
and settings to use for that patient, based on clinical judgment, the condition
and needs of the patient, and the benefits, limitations and characteristics of
the breath delivery options. As the patient's condition changes over time,
periodically assess the chosen modes and settings to determine whether or
not those are best for the patient's current needs.
WARNING:
Avoid nuisance alarms by applying appropriate alarm settings.
WARNING:
To prevent inappropriate ventilation, select the correct Tube Type (ET or
Tracheostomy) and tube inner diameter (ID) for the patients ventilatory
needs. Inappropriate ventilatory support leading to over-or under-ventilation
could result if an ET tube or trach tube setting larger or smaller than the actual
value is entered.
WARNING:
Setting expiratory volume alarms to OFF increases the risk of not detecting a
low returned volume.
WARNING:
Setting any alarm limits to OFF or extreme high or low values can cause the
associated alarm not to activate during ventilation, which reduces its efficacy
for monitoring the patient and alerting the clinician to situations that may
require intervention.

Warnings Regarding Hoses, Tubing, and Accessories


WARNING:
To prevent electrostatic discharge (ESD) and potential fire hazard, do not use
antistatic or electrically conductive hoses or tubing in or near the ventilator
breathing system.

Operators Manual

1-9

Introduction

1-10

WARNING:
Adding accessories to the ventilator can change the pressure gradient across
the ventilator breathing system (VBS) and affect ventilator performance.
Ensure that any changes to the ventilator circuit configurations do not exceed
the specified values for circuit compliance and for inspiratory or expiratory
limb total resistance. Reference Technical Specifications, p. 11-3. If adding
accessories to the patient circuit, always run SST to establish circuit
compliance and resistance prior to ventilating the patient.
WARNING:
Use of a nebulizer or humidifier can lead to an increase in the resistance of
inspiratory and expiratory filters. Monitor the filters frequently for increased
resistance or blockage.
WARNING:
During transport, the use of breathing tubing without the appropriate cuffed
connectors may result in the circuit becoming detached from the ventilator.
WARNING:
The added gas from an external pneumatic nebulizer can adversely affect
spirometry, delivered O2%, delivered tidal volumes, and breath triggering.
Additionally, aerosolized particulates in the ventilator circuit can lead to an
increase in expiratory filter resistance.
WARNING:
Carefully route patient tubing and cabling to reduce the possibility of patient
entanglement or strangulation.
WARNING:
Always use filters designed for use with the Puritan Bennett 980 Series
Ventilator. Do not use filters designed for use with other ventilators.
Reference Covidien Accessories and Options, p. 9-4 for relevant filter part
numbers.
WARNING:
To avoid liquid entering the ventilator, empty the expiratory condensate vial
before fluid reaches the maximum fill line.

Operators Manual

Safety Information

1.3.9

WARNING:
Accessory equipment connected to the analog and digital interfaces must be
certified according to IEC 60601-1. Furthermore, all configurations shall
comply with the system standard IEC 60601-1-1. Any person who connects
additional equipment to the signal input part or signal output part of the
ventilator system configures a medical system, and is therefore responsible
for ensuring the system complies with the requirements of the system
standard IEC 60601-1-1. If in doubt, consult Covidien Technical Services at
1.800.255.6774 or your local representative.

Warnings Regarding Gas Sources


WARNING:
Do not use nitric oxide, helium or mixtures containing helium with the
ventilator. It has not been validated for use with these gas mixtures.
WARNING:
To avoid the risk of ventilator malfunction, do not use the ventilator with
anesthetic gases.
WARNING:
For proper ventilator operation, use only clean, dry, medical grade gases
when ventilating a patient.
WARNING:
Use of only one gas source could lead to loss of ventilation and/or hypoxemia
if that one gas source fails and is not available. Therefore, always connect at
least two gas sources to the ventilator to ensure a constant gas supply is
available to the patient in case one of the gas sources fails. The ventilator has
two connections for gas sources: air inlet, and oxygen inlet.
WARNING:
Use of the ventilator in altitudes higher or barometric pressures lower than
those specified could compromise ventilator operation. Reference
Environmental Specifications, p. 11-7 for a complete list of environmental
specifications.

Operators Manual

1-11

Introduction

WARNING:
The ventilator should be connected to a gas pipeline system compliant to ISO
7396-1:2007 because:

Installation of the ventilator on a non-ISO 7396-1:2007 compliant gas pipeline system may exceed the pipeline design flow capacity.

1.3.10

1-12

The ventilator is a high-flow device and can interfere with the operation
of other equipment using the same gas source if the gas pipeline system
is not compliant to ISO 7396-1:2007.

Warnings Regarding Infection Control


WARNING:
Patients receiving mechanical ventilation may experience increased
vulnerability to the risk of infection. Dirty or contaminated equipment is a
potential source of infection. It is recognized that cleaning, sterilization,
sanitation, and disinfection practices vary widely among health care
institutions. Always follow your hospital infection control guidelines for
handling infectious material. Follow the instructions in this manual and your
institutions protocol for cleaning and sterilizing the ventilator and its
components. Use all cleaning solutions and products with caution. Follow
manufacturers instructions for individual cleaning solutions. Reference
Chapter 7 of this manual.
WARNING:
To prevent infection and contamination, always ensure inspiratory and
expiratory bacteria filters are installed before ventilating the patient.
WARNING:
Never attempt to re-use single patient use components or accessories. Doing
so increases risk of cross-contamination and re-processing of single patient
use components or accessories may compromise functionality leading to
possible loss of ventilation.

Operators Manual

Safety Information

1.3.11

1.3.12

Warnings Regarding Ventilator Maintenance


WARNING:
To ensure proper operation and avoid the possibility of physical injury, this
ventilator should only be serviced by qualified technicians who have received
appropriate Covidien-provided training for the maintenance of this
ventilator.
WARNING:
Follow preventive maintenance according to specified intervals listed in these
tables. Reference Operator Preventive Maintenance Frequency, p. 7-2.
Reference Service Preventive Maintenance Frequency, p. 1-2 in the Operators
Manual Addendum on Preventive Maintenance.

Cautions
Caution:
To prevent possible equipment damage, ensure the casters are locked to
prevent inadvertent movement of the ventilator during routine maintenance,
or when the ventilator is on an incline.
Caution:
Do not use sharp objects to make selections on the display or keyboard.
Caution:
To ensure optimal performance, keep the GUI touch screen and keyboard
clean and free from foreign substances. Reference Surface Cleaning Agents,
p. 7-5.
Caution:
To avoid moisture entering the ventilator and possibly causing a malfunction,
Covidien recommends using a wall air water trap when using piped medical
air from a facility-based air compressor.
Caution:
Use only the cleaning agents specified. Reference Surface Cleaning Agents, p.
7-5. for approved cleaning agents.

Operators Manual

1-13

Introduction

1.3.13

1-14

Caution:
Do not block cooling vents.
Caution:
Ensure proper connection and engagement of expiratory and inspiratory
filters.
Caution:
Follow instructions for proper GUI and BDU (breath delivery unit) mounting
as described in the Puritan Bennett 980 Series Ventilator Installation
Instructions.
Caution:
Follow proper battery installation instructions as described in this manual.
Caution:
When transferring the ventilator from storage conditions, allow its
temperature to stabilize at ambient conditions prior to use.
Caution:
Remove extended and primary batteries from ventilator prior to transporting
in a vehicle. Failure to do so could result in damage to the ventilator.

Notes
Note:
Federal law (USA) restricts the sale of this device except by or on the order of a
physician.
Note:
When using non-invasive ventilation (NIV), the patients actual exhaled volume may
differ from the exhaled volume reported by the ventilator due to leaks around the
mask.

Operators Manual

Safety Information

1.4

1.4.1

Note:
When utilizing a closed-suction catheter system, the suctioning procedure can be
executed using existing mode, breath type, and settings. To reduce potential for
hypoxemia during the procedure, elevated delivered oxygen can be enabled using the
Elevate O2 control. Reference To adjust the amount of elevated O2 delivered for two
minutes, p. 3-40.

Obtaining Technical Assistance


Technical Services
For technical information and assistance, to order parts, or to order an Operators Manual or Service Manual, contact Covidien Technical Services at
1.800.255.6774 or a local Covidien representative.Reference the following
table for service centers in the USA and other countries. The Puritan Bennett
980 Series Ventilator Service Manual includes information necessary to service
or repair the ventilator when used by qualified, factory-trained personnel.
When calling Covidien Technical Services, or a local Covidien representative,
have the BDU and GUI serial numbers available, as well as the firmware version
number of the ventilator system.
The ventilators configuration is available by touching the wrench icon on the
GUI screen, then touching the Options tab. Have this information available
whenever requesting technical assistance.
Information on obtaining the Operators and Service Manuals are available on
the Internet at:
http://www.covidien.com

The following table lists Covidien Service Centers, addresses, telephone, and
Fax numbers:

Operators Manual

1-15

Introduction

1-16

Covidien Argentina
Aguero 351
Capital Federal - 1171
ABC, Argentina
Tel: (5411) 48635300
Fax: (5411) 48634142

Covidien Asia
Singapore Regional
Service Centre
15 Pioneer Hub, #06-04
Singapore 627753
Tel (65) 6578 5288
Fax (65) 6515 5260

Covidien Australia
52A Huntingwood Drive
Huntingwood, NSW
2148
Australia
Tel: (+61) 1800 - 350702
Fax: (+61) 2967 - 18118

Covidien Austria GmbH


Campus 21
Europaring F09402
A-2345 Brunn am Gebirge

Covidien Belgium
BVBA/SPRL.
Generaal De Wittelaan 9/5
2800 Mechelen
Belgium
Tel +32 15 29 44 50
Fax +32 15 29 44 55

Covidien Brazil
Av. Das Naes Undias
12995 Andar 23 - Brooklin
So Paulo, SP
Brasil 04578-000
Tel: (5511) 2187-6200
Fax: (5511) 2187-6380

Covidien Canada
19600 Clark Graham
Baie d'Urfe, QC, H9X
3R8
Canada
Tel:1-514-332-1220,
Select Option 2
Fax: 1-514-695-4965

Covidien Chile
Camino lo Boza (Ex 8395)
Pudehuel
Santiago
Chile
Tel: (562) 739 - 3000
Fax: (562) 783 - 3149

Covidien China
2F, Tyco Plaza
99 Tian Zhou Rd
Shang Hai 200233
P.R. China
Tel: (+86) 4008 1886
86
Fax: (+86) 2154 4511
18

Covidien Colombia
Edificio Prados de la
Morea
Carretera Central Del
Norte
(Cra 7a)Kilometro 18,
Chia-Cundinamarca
Bogota, Colombia
Tel: (571) 619-5469
Fax: (571) 619-5425

Covidien Costa Rica


Global Park, Parkway 50
La Auroa de Heredia
Costa Rica
Tel: (506) 2239 - 5386
Fax: (506) 2239 - 5319

Covidien ECE
Proseck 851/ 64
190 00 Prague
Czech Republc
Tel +42 024 109 57 35
Fax + 42 02 3900 0437

Covidien Danmark
A/S
Langebrogade 6E, 4.
th
DK-1411 Kbenhavn
K
Danmark
Tel +45 4368 2171
Fax:+45 4368 2172

Covidien Deutschland
GmbH
Gewerbepark 1r
D-93333 Neustadt/
Donau
Germany
Tel + 49 (0) 9445 95 9 0
Fax + 49 (0) 9445 95 9
155

Covidien ECE
Galvahiho 7 / A
82104 Bratislava Slovakia
Tel +420 2 41 095 735
Fax +420 2 39 000 437

Covidien Finland Oy
Pursimiehenkatu
26-39C,
PL407
FIN-00151 Helsinki
Finland
Te. +358 9725 192 88
Fax +358 9725 192 89

Covidien France
SAS
2 Rue Denis Diderot
78990 Elancourt
France
Tel +33 (0) 13079 80
00
Fax +33 (0) 130 79 80
30

Covidien Hong Kong


Unit 12 - 16, 18/F
BEA Tower
Millennium City 5
4187 Kwun Tong Road
Kwum Tong,
Kowloon, Hong Kong
Tel + 852 3157 7299
Fax + 852 2838 0749

Covidien India
10th Floor Building No
9B
DLF Cyber City Phase III
Gurgaon
Haryana - 122002
India
Tel + 91 1244 709800
Fax + 91 1244 206850

Covidien ECE
Marissy u.7.
1095 Budapest
Hungary
Tel + 36 1880 7975
Fax + 36 1777 4932

Tel: +43 (0) 2236 37 88 39


Fax: +43 (0) 2236 37 883940

Operators Manual

Safety Information

Covidien Ireland
Block G, Ground
Floor,
Cherrywood Business
Park,
Loughlinstown
County Dublin,
Ireland
Tel +353 (0)
1.4073173
Fax +353(0)
1.4073174

Covidien Israel
5, Shasham St.
North Industrial Park
POB3069
Caesarea, 38900
Tel +972 4.627 73 88
Fax+972 4.627 76 88

Covidien Italia S.p.A


Via Rivoltana 2/D
I-20090 Segrate (Mi)
Italy
Tel +39 02 703 173 1
Fax +39 02 71740584

Covidien Japan Inc.


Technical Support Center
83-1, Takashimadaira 1Chome
Itabashi-ku, Tokyo 175-0082
Japan
Tel: +81 (0) 3 6859 0120
Fax: +81 (0) 3 6859 0142

Covidien Korea
5F, Hibrand Living
Gwan, #215,
Yangjae-Dong,
Seocho-Gu
Seoul, Korea
Tel: +822 570 5459
Fax: +822 570 5499

Covidien Mexico
Insurgentes Sur # 863,
Piso 16
Col. Npoles
Del. Benito Juarez
Mexico, D.F. 03810
Mexico
Tel: (5255) 5804-1524
Fax: (5255) 5536-1326

Covidien Nederland
BV
Hogeweg 105
NL5301 LL
ZaltbommelNederland
Tel0418 57 66 00
Fax 0418 57 67 91

Covidien New Zealand


Cnr Manu Tapu Dr & Joseph
Hammond Pl.
Auckland Airport
New Zealand
Phone: + 64 508 489 264

Covidien Norge AS
Bankveinen 1,
Postboks 343
N-1372 Askerr
Norway
Tel +47 2415 98 87
Fax +47 2415 15 98
88

Covidien Panama
Parque Industrial Costa
del Esta
Calle Primera, Edifio #
109
Panama City, Panama
Tel: (507) 264-7337
Fax: (507) 236-7408

Covidien Polska
Al. Jerozolimskie 162
02-342 Warszawa.
Polska
Tel +48 22 312 20 00
Fax +48 22 312 20 20

Covidien Portugal Lda.


Produtos De Sade Ida.
Est: Outeiro Polima, Lote 101 Piso
Abboda
P-2785-521 S. Domingos de
Rana
Portugal
Tel +351 21 448 10 00
Fax +351 21 445 05 88

Covidien Puerto
Rico
Palmas Industrial Park
Road 869 Km 2.0
Bdlg. #1
Catao, PR 00962
Tel. 787-993-7250
Ext. 7222 & 7221
Fax 787-993-7234

Covidien Russia
53 bld. 5 Dubininskaya
StreetMoscow
RUSSIA. 119054
Tel +70 495 933 64 69
Fax +70 495 933 64 68

Covidien Saglik A.S.


Maslak Mahallesi Bilim
Sokak No: 5, Sun Plaza
Kat: 2-3
Sisli, Istanbul 34398
Turkey
Tel +90 212 366 20 00
Fax +90 212 276 35 25

Covidien South Africa


Corporate Park North
379 Roan Crescent
RandjesparkMidrand, South
Africa
Tel +27 115 429 500
Fax +27 115 429 624

Operators Manual

1-17

Introduction

Covidien Spain S.L.


c/Fructus Gelabert
6, 8a Planta
08970 Sant
JoanDesp
Barcelona, Spain
Tel +34 93475 86 10
Fax +34 93 477 10 17

Covidien Sverige AB
Hemvrnsgatan 9, Box
54
SE-171 74 Solna
Sweden
Tel +46(0)8517 615 73
Fax + 46 (0)8 517 615 79

Covidien UK
4500 Parkway
Whiteley, Fareham
Hampshire
PO157NY, United
Kingdom
Tel +44 (0) 1329
2240002
Fax +44 (0) 1329
220213

Covidien USA
2101 Faraday Ave
Carlsbad, CA 92008
Phone: 1-800-255-6774
(option 4
Email: VentTechSupport@covidien.com

Covidien Switzerland
Roosstrasse 53
Ch-8832 Wollerau
Switzerland
Tel +41(0)44 786 50 50
Fax +41 (0) 44 78650 10

Covidien Thailand
319 Chamchuri Square 17th
Floor, Unit 1-8,
Phayathai Road
Pathumwan, Bangkok
10330, Thailand
Tel +66-2 207-3 100
Fax +66-2 207 - 3101

For online Technical support, visit the SolvITSM Center Knowledge Base at
www.covidien.com. The SolvIT Center provides answers to frequently asked
questions about the ventilator system and other Puritan Bennett products 24
hours a day, seven days a week.
1.4.2

On-Screen Help
The ventilator is equipped with an on-screen help system that enables users to
select an item on the screen and display a description of that item. Follow the
procedure below to access and use on-screen help.
Accessing On-screen Help Topics
Help topics on the ventilator are called tooltips. If a tooltip is available, a glowing
blue outline appears around the item in question.
To access tooltips
1.
Touch the item in question for a period of at least 0.5 s, or drag the help icon (the
question mark icon appearing at the lower right of the GUI screen) to the item in question. A tooltip appears with a short description of the item. Most screen items have
tooltips associated with them, providing the operator with access to a multitude of
help topics.
2.

1-18

Touch more on the dialog to display an expanded description.

Operators Manual

Safety Information

3.

Touch close to close the dialog, or let it fade away after five (5) seconds.

Note:

Dragging the help icon causes the tooltip to display in its unexpanded state.

Dragging the help icon and pausing causes a tooltip to display. Continue dragging
to another item to dismiss the last tooltip and display another tooltip.

Other Resources
Additional resources for information about the ventilator can be found in the
Puritan Bennett 980 Series Ventilator Service Manual and appendices in this
manual for BiLevel 2.0, IE Sync, Leak Sync, PAV+, NeoMode 2.0, and Proximal
Flow Sensor options.
1.5

Warranty Information
To obtain warranty information for a covered product, contact Covidien Technical Services at 1.800.255.6774 or call a local Covidien representative.

1.6

Manufacture Date
The graphical User Interface (GUI) and Breath Delivery Unit (BDU) each possess
a specific year of manufacture applicable only for that assembly. These dates
are contained in the serial numbers for each assembly or option. Serial
numbers for the 980 Ventilator final units consist of ten digits, in the following
format:
35ZYYXXXXX
where

35 signifies the unit was manufactured in Galway, Ireland

Z represents the product code (B= breath delivery unit, G= GUI, P = Proximal Flow
Monitoring option). The product codes shown here are typically the most
common. There may be other product codes shown in the serial number depending upon the particular option(s) purchased.

YY is a two-digit year code that changes with each year

XXXXX is a sequential number that resets at the beginning of each new year

Operators Manual

1-19

Introduction

Serial numbers are located on labels on the back panels of the GUI and BDU,
and in various locations on product options.
1.7

Manufacturer
Covidien llc, 15 Hampshire Street, Mansfield, MA 02048 USA.

1.8

Electromagnetic Compatibility
The ventilator system complies with the requirements of IEC 60601-1-2:2007
(EMC Collateral Standard) including the E-field susceptibility requirements at a
level of 10 volts per meter, at frequencies from 80 MHz to 2.5 GHz. However,
even at this level of device immunity, certain transmitting devices (cellular
phones, walkie-talkies, cordless phones, paging transmitters, RFID devices,
etc.) emit radio frequencies that could interrupt ventilator operation if operated in a range too close to the ventilator. Practitioners should be aware of possible radio frequency interference if portable devices are operated in close
proximity to the ventilator.
The Puritan Bennett 980 ventilator requires special precautions to be taken
regarding electromagnetic compatibility (EMC) and must be installed and put
into service according to the EMC information provided in Chapter 11 in this
manual.

1-20

Operators Manual

2 Product Overview

2.1

Overview
This chapter contains introductory information for the Puritan Bennett 980
Series Ventilator.

Note:
Items shown in bold-italic font are contained as entries in the glossary.

Communication between the ventilators graphical user interface (GUI) and


the breath delivery unit (BDU) occurs continuously via independent central
processing units (CPUs).
Reference Pneumatic Diagram, p. 2-37 and its associated reference designators when reading the following paragraphs.
Gas delivery starts with the ventilator connected to wall (or bottled) air and
oxygen. Gas travels to the mix module where gas pressures are regulated by
their respective proportional solenoid valves (PSOLs). The PSOLs meter the
gases according to the ventilator settings entered, then the gases flow through
individual air and oxygen flow sensors into the mix manifold and accumulator
for mixing. The individual gas pressures are continuously monitored both
before and after they are mixed in the mix manifold and accumulator assemblies. The mixed gas then flows to the inspiratory pneumatic system where it
flows through the breath delivery flow sensor and then the inspiratory PSOL
for delivery to the patient.
Before the gas reaches the patient, it passes through an internal inspiratory
bacteria filter, then through an external inspiratory bacteria filter attached to
the ventilators gas outlet (the To patient port) where the breathing circuit is
attached. When the gas returns from the patient, it flows through the expiratory limb of the breathing circuit, to the From patient port on the expiratory
bacteria filter (which includes a condensate vial) before flowing through the
exhalation valve flow sensor and exhalation valve (EV). A gas exhaust port
allows exhaled gas to exit the ventilator and flow to the room.

2-1

Product Overview

The ventilator recognizes the patients breathing effort using pressure triggering (P-TRIG), flow triggering (V-TRIG) or IE Sync triggering (if the IE Sync Option
is installed).Reference Appendix C for more information on the IE Sync Option.
During pressure triggering, as the patient inhales, the airway pressure decreases and the inspiratory pressure transducer (PI) monitors this pressure
decrease.When the pressure drops to at least the value of the pressure sensitivity (PSENS) setting, the ventilator delivers a breath. During flow triggering,
the difference between inspiratory and expiratory flows is monitored. As the
patient inhales, the exhalation valve flow sensor measures less flow, while the
delivery flow sensor measurement remains constant. When the difference
between the two measurements is at least the value of the operator-set flow
sensitivity (VSENS), the ventilator delivers a breath. If the patient is not inhaling,
any difference between delivered flow and expiratory flow is due to flow
sensor inaccuracy or leaks in the ventilator breathing circuit. To compensate for
leaks, which can cause autotriggering, the clinician can increase the VSENS
setting or enable Leak Sync, if available.

Note:
Leak Sync is a software option. Details on its operation are provided in the Leak Sync
appendix in this manual. Reference Appendix B.

A backup pressure triggering threshold of 2 cmH2O is also in effect. This provides enough pressure sensitivity to avoid autotriggering, but will still allow the
ventilator to trigger with acceptable patient effort.
The exhalation valve controls Positive End Expiratory Pressure (PEEP) using
feedback from the expiratory pressure transducer (PE). The valve controller
also cycles the ventilator into the exhalation phase if the PE measurement
equals or exceeds the operator-set high circuit pressure limit. The PE measurement also controls when the safety valve (SV) opens. If PE measures 110
cmH2O or more in the ventilator breathing circuit, the safety valve opens,
allowing the patient to breathe room air (if able to do so) through the valve.
2.2

Ventilator Description
The ventilator system is available in three models. All ventilators provide continuous ventilation to patients requiring respiratory support.

2-2

Puritan Bennett 980 Pediatric - Adult Ventilator The Pediatric - Adult


model ventilates Pediatric or Adult patients with predicted body weights from 3.5
kg to 150 kg, and with tidal volumes from 25 mL to 2500 mL.

Operators Manual

GUI Controls and Indicators

Puritan Bennett 980 Neonatal Ventilator The Neonatal model ventilates Neonatal patients with predicted body weights from 0.3 kg to 7.0 kg, and
with tidal volumes for mandatory volume-controlled breaths from 2 mL to 320
mL.

Puritan Bennett 980 Universal Ventilator The Universal model ventilates


Neonatal, Pediatric, and Adult patients with predicted body weights from 0.3 kg
to 150 kg, and with tidal volumes for mandatory volume-controlled breaths from
2 mL to 2500 mL.

To ventilate Neonatal patients on the Pediatric - Adult or Universal models, the


NeoMode 2.0 software option is required. For details regarding the NeoMode
2.0 software option, Reference Appendix E in this manual.
The estimated service life of the ventilator is approximately 10 years, provided
the preventive maintenance schedule stated in the Puritan Bennett 980
Series Ventilator Service Manual is followed; however, service life of individual
units may vary.
The ventilators IEC 60601-1/EN 60601-1 classification is:

Protection class I

Type BF

Mobile

Internally powered

IP 21 equipment

Continuous operation

Not suitable for use with flammable medical gases (not AP or APG)

Reference BDU Rear Label or Panel Symbols and Descriptions, p. 2-11 for a
description of the meaning of the IP classification.
The ventilator system uses a graphical user interface (GUI) and breath
delivery unit (BDU) for entering patient settings and delivering breaths to the
patient. The GUI contains electronics capable of transferring the clinicians
input (by touching the screen) to the BDU where pneumatic and electronic systems, respectively, generate the breathing parameters.

Operators Manual

2-3

Product Overview

2.3

Indications For Use


The Puritan Bennett 980 Ventilator System is designed for use on patient
population sizes from Neonatal (NICU) through Adult who require respiratory
support or mechanical ventilation and weigh a minimum of 0.3 kg (0.66 lb). It
is suitable for service in hospital (institutions) and intra-hospital transport to
provide continuous positive pressure ventilatory support, delivered invasively or
noninvasively, to patients who require the following types of ventilator support:

Positive Pressure Ventilation, delivered invasively (via endotracheal tube or trach


tube) or non-invasively (via mask or nasal prongs)

Assist/ Control, SIMV or Spontaneous modes of ventilation

Note:
Intended typical usage may be defined to include the following for the ventilator
system:
Hospital Use Typically covers areas such as general care floors (GCFs), operating
rooms, special procedure areas, intensive and critical care areas within the hospital
and in hospital-type facilities. Hospital-type facilities include physician office-base
facilities, sleep labs, skilled nursing facilities, surgicenters, and sub-acute centers.
Intra-hospital transport Includes transport of a patient within the hospital or
hospital-type facility. All external hospital transportation (i.e. ambulance or aircraft) is excluded.

2.4

Note:
Federal law (USA) restricts the sale of this device except by or on the order of a
physician.

Contraindications
Do not operate the ventilator in a magnetic resonance imaging (MRI) environment.

2-4

Operators Manual

GUI Controls and Indicators

2.5

Components List
Note:
No parts of the ventilator system contain latex.
Note:
The components in the gas pathway that can become contaminated with bodily fluids
or expired gases during both normal and single fault conditions are:

External inspiratory filter

Internal inspiratory filter

Expiratory filter and condensate vial

Exhalation valve assembly

The typical ventilator system ships with the following packing list. Depending
upon the ventilator system purchased, your list may vary.
Table2-1.Typical Packing List
Quantity

Operators Manual

Item

Graphical User Interface

Breath Delivery Unit

Inspiratory filter

Expiratory filter

Condensate vial

Gas hoses (air and oxygen)

Standard caster base

Power cord

Operators Manual CD

Puritan Bennett 980 Series Ventilator Installation Instructions

Flex arm

Drain bag

Gold standard circuit (for running EST)

2-5

Product Overview

2.6

2.6.1

Product Views
GUI Front View
Figure2-1.GUI Front View

2-6

Display brightness key

Inspiratory pause key

Display lock key

Expiratory pause key

Alarm volume key

Alarm reset key

Manual Inspiration key

Alarm silence key

Rotary encoder (knob)

Operators Manual

GUI Controls and Indicators

2.6.2

GUI Rear View


Figure2-2.GUI Rear View

Reference Common Symbols Found on GUI or BDU Labels, p. 2-13 for symbols
found on the GUI or BDU. The Do Not Push symbol found on the GUI, only,
is shown in this table.

Operators Manual

2-7

Product Overview

2.6.3

BDU Front View


Figure2-3.BDU Front View

2-8

Condensate vial

Power switch

Expiratory filter

Status display

Expiratory filter latch

Internal inspiratory filter

AC power indicator

Option connector panel door

Operators Manual

GUI Controls and Indicators

Table2-2.BDU Front Label Symbols and Descriptions


Symbol

Description
To Patient port

From Patient port

Expiratory filter latch locked (down)/unlocked (up)

Operators Manual

2-9

Product Overview

2.6.4

BDU Rear View


Figure2-4.BDU Rear View

Standard base

Service mode button

Air inlet

Remote alarm port

Oxygen inlet

Cylinder mount (optional)

Labels indicating installed software


options

Software option labels are applied to the grid located on the back of the ventilator, as shown below and in the previous image (item 4).

2-10

Operators Manual

GUI Controls and Indicators

Figure2-5.Installed Software Options

The following table lists the symbols and descriptions found on BDU or base
labels.
Table2-3.BDU Rear Label or Panel Symbols and Descriptions
Symbol

Description
US federal law restricts this device to sale by or on the order of a physician.
User must consult instructions for use. Symbol is also found on Do not
obstruct labels on both left and right sides of the ventilator, and on
label indication supply gas connections.
Keep away from fire or flame. Oxygen rich environments accelerate
combustibility.

Atmospheric pressure limitations The operational atmospheric pressure range 70 kPa to 106 kPa (10.2 psi to 15.4 psi).

Humidity limitations The operational humidity limit range 10% to


95%.

Temperature limitations The operational temperature limit range


50F to 104F (10C to 40C).

Operators Manual

2-11

Product Overview

Table2-3.BDU Rear Label or Panel Symbols and Descriptions (Continued)


Symbol

Description
Type BF applied part.

IEC Ingress protection classification Protected against ingress of


fingers or similar objects and protected from condensation.

Explosive hazard. Do not use in the presence of flammable gases.

Authorized to bear the CSA certification mark signifying the product


has been evaluated to the applicable ANSI/Underwriters Laboratories
Inc. (UL) and CSA standards for use in the US and Canada.

The ventilator contains components manufactured with phthalates.

Unsafe to use the ventilator in magnetic resonance imaging environments.

Potential equalization point (ground) (on AC panel).

CB1

BDU Circuit Breaker (on AC panel).

CB2

Compressor Circuit Breaker (on AC panel).


USB port (at rear of ventilator).

HDMI port (at rear of ventilator).

Service port (at rear of ventilator).

2-12

Operators Manual

GUI Controls and Indicators

Table2-3.BDU Rear Label or Panel Symbols and Descriptions (Continued)


Symbol

Description
Service mode button (at rear of ventilator).

Remote alarm port (at rear of ventilator).

Ethernet connector (at rear of ventilator).

Serial port (at rear of ventilator).

Table2-4.Common Symbols Found on GUI or BDU Labels


Symbol

Description
Do Not Push - Do not push on the GUI

Manufacturer - Name of the ventilator manufacturer.

Serial number.
Manufacture date The manufacture date is contained in the serial
number. Reference Manufacture Date, p. 1-19 for details regarding
interpretation of the serial number.
WEEE Proper waste disposal. Follow local governing ordinances
regarding disposing of waste labeled with the WEEE symbol.

Operators Manual

2-13

Product Overview

2.6.5

Ventilator Side Views


Figure2-6.Ventilator Right Side View

2-14

Operators Manual

GUI Controls and Indicators

Figure2-7.Ventilator Left Side View

2.7

Mounting Configurations
The ventilator system can be mounted as a free-standing unit standing at the
patients bedside; the BDU with the GUI is mounted on a base with casters and
includes a handle for ease of movement.

2.8

Battery Backup
The ventilator system uses a battery to provide backup power in case AC
power is lost. When operating on battery power, the status display shows the
On Battery Power image, and the GUI displays a representation of battery
charge levels. Reference Typical Status Display Indicators and Messages, p. 230 to for a description of the status display images and messages. An optional,
extended battery is available to lengthen the amount of time the ventilator can
operate on battery power. Reference Using Battery Power, p. 3-3.

Operators Manual

2-15

Product Overview

2.9

Graphical User Interface


There are two displays on the ventilator the primary display (GUI) and the
status display.

2.9.1

Primary Display
The GUI incorporates a 15 display that rotates throughout a 170 angle about
a vertical axis in either direction. The GUI can also be tilted up to 45 from vertical.
The clinician enters ventilation parameters via the GUIs touch screen, also
known as the ventilators primary display. The GUIs keys activate other ventilator functions including screen brightness, display lock, alarm volume, manual
inspiration, inspiratory pause, expiratory pause, alarm reset, and alarm silence.
The GUI displays the following information depending on the state of the ventilator:

2.10

2.10.1

Ventilator, apnea, and alarm settings

Patient data

Waveforms

Current alarm banners

GUI Controls and Indicators


Control Keys
The GUI bezel has eight off-screen control keys as shown below.
Table2-5.GUI Control Keys

Key symbol

Description
Brightness control key Adjusts the GUI screen brightness. Press the key and turn
the knob to adjust the brightness.

2-16

Operators Manual

GUI Controls and Indicators

Table2-5.GUI Control Keys (Continued)


Key symbol

Description
Display lock key Actuates a lock to prevent inadvertent settings changes to the
ventilator (including the knob function) while the display is locked. The display lock
is useful when cleaning the touch screen. Press the key again to unlock the display.
Alarm volume key Adjusts the alarm volume. The alarm volume cannot be
turned OFF.

Manual inspiration key In A/C, SIMV, and SPONT modes, delivers one manual
breath to the patient in accordance with the current mandatory breath parameters.
In BiLevel mode, transitions from low pressure (PL) to high pressure (PH) (or vice
versa). To avoid breath stacking, a manual inspiration is not delivered during inspiration or during the restricted phase of exhalation. Reference Manual Inspiration,
p. 10-22 for information on the restricted phase of exhalation.
The Manual inspiration key can be used to deliver mandatory breaths to the patient
or to run an inspiratory pause maneuver in SPONT mode. The manual inspiration
key cannot be used to run an expiratory pause maneuver in SPONT mode.
Inspiratory pause key Initiates an inspiratory pause which closes the inspiratory
and exhalation valves and extends the inspiratory phase of a mandatory breath for
the purposes of measuring end inspiratory pressure (PI END) for calculation of
plateau pressure (PPL), static compliance (CSTAT), and static resistance (RSTAT).
Expiratory pause key Initiates an expiratory pause which extends the expiratory
phase of the current breath in order to measure total PEEP (PEEPTOT).

Alarm reset key Clears active alarms or resets high-priority alarms and cancels an
active alarm silence. An alarm reset is recorded in the alarm log if there is an active
alarm. DEVICE ALERT alarms cannot be reset.
Alarm silence key Silences alarms for two minutes. Cancel the alarm silence
function by touching the on-screen Cancel button.

2.10.2

Visual Indicators
The table below shows the GUIs visual indicators. Reference Areas of the GUI,
p. 4-3 for area names.
The alarm silence function has two visual indicators the alarm silence key on
the GUI bezel glows yellow during an alarm silence interval, and a visual count-

Operators Manual

2-17

Product Overview

down timer appears, showing the amount of time the alarm silence interval
has remaining.
Table2-6.GUI Visual Indicators
Symbol

Description
Ventilator Setup (Vent Setup) button. Located at
the lower left corner of the GUI. Touch this
button to open the ventilator setup screen.

Adult patient circuit indicator. Indicates adult


circuit type tested during SST, and in use.
Appears above the Vent Setup button.

Pediatric patient circuit indicator. Indicates pediatric circuit type tested during SST, and in
use.Appears above the Vent Setup button.

Neonatal patient circuit indicator. Indicates neonatal circuit type tested during SST, and in
use.Appears above the Vent Setup button.

Home icon. A constant access icon. Reference


Areas of the GUI, p. 4-3. Touch this icon to
dismiss all open dialogs on the GUI screen. The
display resumes showing the ventilator waveforms.
Alarms icon. A constant access icon. Reference
Areas of the GUI, p. 4-3.Touch this icon to display
the alarm settings screen, which allows alarm
limits to be changed.
Logs icon. A constant access icon. Reference
Areas of the GUI, p. 4-3. Touch this icon to
display the logs screen, which contains tabs for
Alarms, Settings, Patient Data, Diagnostics, EST/
SST status, General Event, and Service logs.

2-18

Operators Manual

GUI Controls and Indicators

Table2-6.GUI Visual Indicators (Continued)


Symbol

Description
Elevate O2 control. A constant access icon. Reference Areas of the GUI, p. 4-3. Touch this icon to
increase the set the elevated oxygen concentration to the institutional default O2 configuration
(if institutional default has been configured) for
two minutes, or allows the operator to determine
the additional percentage of oxygen to increase.
The O2 concentration for the two-minute
increase can be set to any value between 1% and
100% O2. If the Elevate O2 function is active,
touching Extend re-starts the two-minute interval. The Elevate O2 function can be terminated
prior to completion of the two-minute interval by
touching Stop. Any time the Elevate O2 control is
activated, an entry is made to the patient data
log.
Screen capture icon. A constant access icon. Reference Areas of the GUI, p. 4-3. Touch this icon
to capture the image displayed on the GUI
screen. Reference To capture GUI screens, p. 5-2
to read the complete procedure for capturing
screen images.
Help icon. A constant access icon. Reference
Areas of the GUI, p. 4-3. Drag this icon to the
item in question and release. A tooltip will appear
describing the items function.
Unread items icon. When this icon appears overlaid on another icon or tab (the logs icon, for
example) it indicates there are unread items at
this location.
Configure icon. A constant access icon. Reference Areas of the GUI, p. 4-3. Touch this icon to
display the configure screen. Tabs with SST
results, options, Comm setup, and date/ time
change are displayed.
Pause icon. Located above the constant access
icons. Touch this icon to pause the waveform
graph.
Waveform layout icon. Located above the constant access icons area.Touch this icon to open
the waveform layout dialog.

Operators Manual

2-19

Product Overview

Table2-6.GUI Visual Indicators (Continued)


Symbol

Description
Grid lines icon. Located above the constant
access icons area. Touch this icon to turn waveform grid lines ON or OFF.
Maximize waveform icon. Located at the upper
right portion of each waveform. Touch this icon
to enlarge the waveform to its maximum size.
Restore waveform icon. Restores waveform to its
original size. Located at the upper right of the
maximized waveform.
Pushpin icon pinned state. When in the pinned
state, prevents a dialog from closing (under
certain conditions). Located in the upper right
corner of the GUI on the vent setup screen. Reference Pushpin Icon, p. 4-5.
Pushpin icon unpinned state. When the
unpinned icon is touched, the pinned state
becomes active. Located in the upper right corner
of the GUI on the vent setup screen. Reference
Pushpin Icon, p. 4-5.
Low priority alarm icon (appears on alarm banner).
Medium priority alarm icon (appears on alarm
banner).
High priority alarm icon (appears on alarm banner).

2.10.3

On-screen Symbols and Abbreviations


Touch an on-screen symbol briefly (0.5 s) to display a tooltip on the GUI screen.
The tooltip contains a definition of the symbol and other descriptive text, available with either short or long descriptions. The short description expands to
show more information by touching more on the tooltip dialog or collapses
by touching less. The tooltip closes by touching close or fades in five (5)
seconds if left alone. Expanding the tooltip dialog prevents the tooltip from
timing out. Touching outside the tooltip causes the dialog to close.
The table below summarizes the ventilators symbols and abbreviations.

2-20

Operators Manual

GUI Controls and Indicators

Table2-7.Symbols and Abbreviations


Symbol or Abbreviation

Definition

TA

Apnea interval

DSENS

Disconnect sensitivity

CDYN

Dynamic compliance

RDYN

Dynamic resistance

EEF

End expiratory flow

PI END

End inspiratory pressure

LEAK

Exhalation leak

VTE MAND

Exhaled mandatory tidal volume

VE TOT

Exhaled minute volume

VE SPONT

Exhaled spontaneous minute volume

VTE SPONT

Exhaled spontaneous tidal volume

VTE

Exhaled tidal volume

ESENS

Expiratory sensitivity

TE

Expiratory time
Flow pattern (ramp)
Flow pattern (square)

Operators Manual

VSENS

Flow sensitivity

V-TRIG

Flow triggering

PH

High pressure setting (in BiLevel)

TH

High pressure time (in BiLevel)

TH:TL

High pressure time to Low pressure time ratio (in BiLevel)

I:E

Inspiratory Time to Expiratory Time (I:E)

C20/C

Inspiration compliance ratio

VLEAK

Inspiratory leak

TI

Inspiratory time

2-21

Product Overview

Table2-7.Symbols and Abbreviations (Continued)

2-22

Symbol or Abbreviation

Definition

PI

Inspiratory pressure

VTI

Inspired tidal volume

VTL

Inspired tidal volume (when Leak Sync is enabled)

PEEPI

Intrinsic PEEP (auto PEEP)

PEEPI PAV

PAV-based intrinsic PEEP

PL

Low pressure setting (in BiLevel)

TL

Low pressure time (in BiLevel)

PMEAN

Mean circuit pressure

NIF

Negative inspiratory force

O2%

Oxygen percentage

P0.1

Airway occlusion pressure at 100 ms

CPAV

PAV-based lung compliance

EPAV

PAV-based lung elastance

% Supp

Percent support setting for Tube Compensation and PAV+

RPAV

PAV-based patient resistance

RTOT

PAV-based total airway resistance

WOBTOT

PAV-based work of breathing of patient and ventilator


during inspiration

PPEAK

Peak circuit pressure

PEF

Peak expiratory flow

VMAX

Peak inspiratory flow

PSF

Peak spontaneous flow

PEEP

Set or monitored positive end expiratory pressure

%Leak

Percent leak

PPL

Plateau pressure

TPL

Plateau time

PCOMP

Compensation pressure

Operators Manual

GUI Controls and Indicators

Table2-7.Symbols and Abbreviations (Continued)


Symbol or Abbreviation

Definition

PSENS

Pressure sensitivity

PSUPP

Pressure support level

P-TRIG

Pressure triggering

VTIY

Proximal inspired tidal volume

VTEY

Proximal exhaled tidal volume

VTI MANDY

Proximal mandatory inspired tidal volume

VTI SPONTY

Proximal spontaneous inspired tidal volume

VTLY

Proximal inspired tidal volume with Leak Sync enabled

Respiratory rate or apnea respiratory rate


Rise time percent

2.10.4

f/VT

Spontaneous rapid/shallow breathing index

TI SPONT

Spontaneous inspiratory time

TI/TTOT

Spontaneous inspiratory time ratio

CSTAT

Static compliance

RSTAT

Static Resistance

VT

Tidal Volume

PEEPTOT

Total PEEP

fTOT

Total Respiratory Rate (monitored)

VC

Vital Capacity

VS

Volume support

Audible Indicators
A tone sounds when a button on the GUI is touched, and also when settings
are accepted. Audible indicators include pitched tones, beeps, and key clicks.
Key clicks sound whenever a key on the GUI is pressed. Various tones annunciate patient alarms.

Operators Manual

2-23

Product Overview

Note:
Pressing ALARM SILENCE mutes alarms for the two-minute alarm silence duration
period.

Caregivers may choose to silence alarms by pressing the ALARM SILENCE key.
A two-minute countdown timer appears on the GUI during the alarm silence
interval. Cancel the alarm silence function by touching Cancel.
Click each icon below to listen to a sample of the corresponding tones:

Note:
To hear the tones, Adobe Acrobat Reader version 10 or higher must be installed on
your computer. Get Adobe Acrobat Reader, free, here.
Table2-8.GUI Audible Indicator Functions
Function

2-24

Description

Low priority alarm tone

A series of two tones. Sounds when a low priority


alarm occurs.

Medium priority alarm tone

A repeating series of three tones. Sounds when a


medium priority alarm occurs.

High priority alarm tone

A repeating series of five tones. Sounds when a


high priority alarm occurs.

Soft bound tone

One tone. Sounds when a soft bound is reached


when making changes to ventilator settings. A
soft bound is a selected value that exceeds or
goes below its limit and requires acknowledgment to continue.

Hard bound tone (invalid entry)

The invalid entry sound occurs when a hard


bound is reached when making changes to ventilator settings. A hard bound defines the upper
or lower limit of the setting, where the setting
cannot be adjusted higher or lower.

Operators Manual

GUI Controls and Indicators

The clinician enters ventilation parameters via the GUIs touch screen. Reference GUI Front View, p. 2-6. The keys activate other ventilator functions. Reference GUI Control Keys, p. 2-16.
2.11

Breath Delivery Unit


The breath delivery unit contains the hardware and software to enable the ventilator to provide patient support.

2.11.1

BDU Controls and Indicators


BDU Controls

Operators Manual

ON/OFF switch Lift the switch cover and turn the ventilator ON or OFF.

2-25

Product Overview

Figure2-8.Ventilator Power Switch and AC Indicator

2-26

AC power indicator

ON/OFF switch

Service mode button Press and release this button when the Covidien splash
screen appears on the status display after powering on the ventilator to enter
Service mode. Reference Service Mode Button (TEST), p. 2-27. Reference Status
Display Indicators and Descriptions, p. 2-30 for an image of the splash screen.

Operators Manual

GUI Controls and Indicators

Figure2-9.Service Mode Button (TEST)

Service mode button

Note:
The Covidien splash screen shows the Covidien logo and appears momentarily as a
banner on the status display.

BDU AC Indicator
The status display and the AC power indicator are the only visual indicators on
the BDU. The AC indicator illuminates green whenever the ventilator is connected to AC power. All other visual indicators on the ventilator are on the
GUI. Reference Typical Status Display Indicators and Messages, p. 2-30 for a
description of the status display indicators and symbols. Reference the section
below for a summary of the information appearing on the status display.

Operators Manual

2-27

Product Overview

Status Display
The status display is a separate display located on the BDU. Reference BDU
Front View, p. 2-8, item 6. The status display provides the following information according to the state of the ventilator:
During normal ventilation the status display shows

Current power source (AC or DC)

Safe State status: (Safety Valve Open (SVO) or Vent Inop

Presence of primary and extended batteries and their charging status

Relative available battery charge level

Circuit pressure graph displaying pressure units, 2PPEAK alarm setting and current
PPEAK and PEEP values

Connection of air and oxygen

Ventilator operational hours

Visual indication of current alarm volume setting

Note:
The status display provides a redundant check of ventilator operation.If the GUI stops
operating for any reason, ventilation continues as set.

The figure below shows a sample of the status display during normal ventilation.

2-28

Operators Manual

GUI Controls and Indicators

Figure2-10.Sample Status Display During Normal Ventilation

Primary and extended battery status


(presence or absence).

PPEAK alarm setting

Alarm volume setting

Measured inspiratory pressure (changes


as pressure changes)

Gas connection status

Selected pressure units

Power status

Measured PEEP

Measured peak circuit pressure (updated at the end of the current breath)

During Service mode the status display supplies

Ventilator serial number(s)

Ventilator operational time

EST and SST history

Power On Self Test (POST) status

Hours until next preventive maintenance is due

Gas pressure at the manifold inlets

Operators Manual

2-29

Product Overview

Reference the table below for status display possibilities.


Typical Status Display Indicators and Messages
The following table lists indicators and messages that appear on the status display:
Table2-9.Status Display Indicators and Descriptions
Status Display Indicator or Message

Meaning
Splash screen. Appears when the ventilators
power switch is turned on. When this image
appears, press and release the TEST button at the
back of the ventilator to enter Service mode.

POST failure. This image appears if a POST error


occurs at ventilator start-up, along with the error
code (in this case a missing primary battery).

Prior to patient connection. The status display


appears as shown when the patient has not been
connected to the ventilator. Note the absence of
PPEAK and PEEP values.

2-30

Operators Manual

GUI Controls and Indicators

Table2-9.Status Display Indicators and Descriptions (Continued)


Status Display Indicator or Message

Meaning
Stand-by state. The status display appears as
shown when the ventilator is in stand-by state.

Battery charged. The ventilators primary battery


(in the right-most slot) is shown fully charged.
The percentage indicator shows 100%.

Battery charging/discharging. Identifies that the


ventilators primary battery and/or extended
battery is charging or discharging. This icon is
animated; orange bars scroll upward indicating
the battery is charging and downward when the
battery is discharging. If the orange bars are
absent, the battery is neither charging nor discharging. A percentage indicator shows the
remaining battery capacity. If an extended
battery is installed, the image shows a similar representation in the extended battery location (leftmost slot).
Battery icon. Denotes the ventilator is operating
on battery power when this image appears on
any status display indicator. Alerts the operator
there is insufficient AC power to operate the ventilator. The indicator is replaced by the on AC
power indicator when adequate AC power is
restored.

Operators Manual

2-31

Product Overview

Table2-9.Status Display Indicators and Descriptions (Continued)


Status Display Indicator or Message

Meaning
On battery power. Alerts the operator there is
insufficient AC power to operate the ventilator.
Ventilator is operating on battery power with
greater than ten minutes of capacity remaining.
Note the appearance of the battery icon.

Low battery. Identifies that the ventilators


primary battery (right-most slot) is discharging
and there are ten minutes or less of battery
capacity remaining. A percentage indicator
shows the remaining battery capacity. If an
extended battery is installed, the image would
show a similar representation in the extended
battery location (left-most slot).

Critically low battery. Identifies that the ventilators primary battery has less than five minutes of
battery capacity remaining. A percentage indicator shows the remaining battery capacity. If an
extended battery is installed, the image would
show a similar representation in the extended
battery location.

Power failure. Alerts the user that the ventilators


battery is depleted or depletion is imminent.
Replace primary or extended battery with a fully
charged batter or connect ventilator to AC
power.

2-32

Operators Manual

GUI Controls and Indicators

Table2-9.Status Display Indicators and Descriptions (Continued)


Status Display Indicator or Message

Meaning
Battery inoperative. This image appears on the
status display when a battery fault renders the
battery inoperative.

Battery not installed. This image appears when


there is no primary battery installed, and renders
ventilator inoperative. This image displays when
the primary battery is removed during ventilator
operation.

Loss of GUI. Indicates a loss of communication


between the BDU and the GUI, where the GUI
cannot reliably display or receive ventilator information. This loss of communication could be
caused by problems such as a GUI hardware failure, faulty Ethernet, or GUI reset. If the indicator
is visible, ventilation continues as currently set.
Replace the ventilator as soon as it is appropriate
to do so. Service the ventilator prior to returning
it for use on patients.
Ventilator inoperative (Vent Inop). Indicates the
ventilator is no longer capable of ventilating a
patient and requires service. The alarm reset key
cannot be used to restore function to the ventilator during a ventilator inoperative condition.
Provide alternate means of ventilation immediately. Note the display of the Safety Valve Open
indicator.

Operators Manual

2-33

Product Overview

Table2-9.Status Display Indicators and Descriptions (Continued)


Status Display Indicator or Message

Meaning
Safety Valve Open (SVO) indicator. During SVO,
the patient can breathe room air through the
safety valve, to the extent the patient is able to
breathe unaided. Reference Safety Valve Open
(SVO), p. 4-36 for more information on the SVO
state.

On AC power indicator. When this image


appears on any status display indicator, indicates
the ventilator is operating on AC power.

Status display appearance when the ventilator is


breathing in Normal mode. Note the appearance
of the AC power icon.

Air available indicator. When this image appears


on any status display indicator, indicates the ventilator is connected to a pressurized air source.

O2 available indicator. When this image appears


on any status display indicator, indicates the ventilator is connected to a pressurized O2 source.

2-34

Operators Manual

GUI Controls and Indicators

BDU Audible Indicators


The continuous tone alarm is the only audible indicator in the BDU, and is
described in the table, below.
Table2-10.BDU Audible Indicator Functions
Indicator

Description

Continuous tone alarm


(Immediate priority)

A continuous tone annunciated when there is a Ventilator Inoperative (Vent Inop)


condition. This alarm lasts for a minimum of two (2) minutes.

2.11.2

Connectors
The ventilator incorporates the following connectors:

2.12

Ventilator outlet port (To patient) A coaxial 15 mm (ID) / 22 mm (OD)


conical connection to which the external inspiratory bacteria filter attaches.

Exhalation port (From patient) port The expiratory limb of the patient
circuit attaches to the inlet of the expiratory bacteria filter. This port is compatible
with a standard 22mm (OD) conical connection.

Proximal Flow sensor A keyed pneumatic connector for the Proximal Flow
Sensor is provided with a locking feature to prevent inadvertent disconnection.
The proximal flow sensor measures flow and pressure at the patient wye. The
Proximal Flow Sensor is an optional sensor. Details on operation are provided in
the appendix in this manual. Reference Appendix F.

Standard interface connectors USB, HDMI, and Ethernet connectors are


provided. The USB connector allows images to be captured on an external USB
storage device and allows communication with an external patient monitor via
serial-over-USB protocol, and the HDMI connector allows the GUI image to be displayed on an external video display device. The Ethernet connector is used by
Service Personnel to upload new software and options. Reference Port Use, p. 519 for more information. Reference To configure Comm ports, p. 5-4 for information on serial-over-USB data transfer when configuring Comm ports for external
devices.

Special Features
A Proximal Flow option is available. The proximal Flow Sensor is used to
measure low flows and pressures associated with neonatal ventilation. If the
ventilator is configured with this option, Reference Appendix F for more information.

Operators Manual

2-35

Product Overview

2.13

Color Definitions
Reference the following figures to view the ventilators pneumatic diagram
during inspiration with various colors representing the gases as shown below.

Table2-11.Color Legend
Color or
Symbol

Description
High-pressure Oxygen (NFPA 99 designation)

High-pressure Air (NFPA 99 designation)

Mixed gases, including air

Atmosphere

Vacuum

Water

2.14

Pneumatic Diagrams
The following figures illustrate the ventilators pneumatics with and without
the optional Proximal Flow System. The Proximal Flow System is only for use
with neonatal patients.

2-36

Operators Manual

GUI Controls and Indicators

Figure2-11.Pneumatic Diagram

Operators Manual

Pressure switch, mix accumulator (PS1)

12

Pressure Sensor, barometric (PA)

Solenoid Valve, options supply (SOL2)

13

Vial, exhalation condensate (ECV)

Pressure sensor, mix accumulator (PMX)

14

Filter, expiratory (F4)

Accumulator, mix (ACCM)

15

Flow sensor assembly, exhalation valve

Tube, mix (TM)

16

Exhalation valve (EV)

Proportional solenoid valve, patient gas


delivery (PSOLD)

17

Filter, exhalation pressure line (F5)

Solenoid valve, BUV (SOL3)

18

Solenoid valve, exhalation pressure


autozero (SOL5)

Safety valve (SV)

19

Pressure sensor, exhalation (PE)

Pressure Sensor, safety valve (PSV)

20

Humidifier

10

Solenoid valve, inspiratory pressure


sensor autozero (SOL4)

21

Filter, External bacterial (FD2)

11

Pressure Sensor, inspiratory (PI)

22

Filter, Internal bacterial (FD1)

2-37

Product Overview

23

Check valve, patient gas delivery (CVD)

34

Check valve, compressor air inlet


(CVCAir)

24

Sensor, oxygen (OS)

35

Filter bowl assembly, Air (WT2)

25

Restrictor, breath delivery bypass (R2)

36

Filter element, Air (F2)

26

Flow sensor, patient gas delivery (FSD)

37

Check valve, wall Air inlet (CVWAir)

27

Check valve, Oxygen (CVO2)

38

Filter, Oxygen Impact (F1)

28

Check valve, Air (CVAir)

39

Filter element, Oxygen (F3)

29

Proportional solenoid valve, Oxygen


(PSOLO2)

40

Pressure sensor, Oxygen gas inlet (PO2)

30

Flow sensor, Air (FSAir)

41

Flow sensor, Oxygen (FSO2)

31

Proportional solenoid valve, Air


(PSOLAir)

42

Restrictor, Prox Flow (R4)

32

Pressure sensor, air gas inlet (PAir)

43

Relief Valve, mix accumulator (RVMA)

33

Restrictor, Wall air bleed outlet (R1)

44

Solenoid Valve, mix accumulator purge


(SOL1)

Figure2-12.Pneumatic Diagram Optional Proximal Flow System

2-38

Operators Manual

GUI Controls and Indicators

Note:
Items enclosed by dotted line represent components internal to the ventilator.

Operators Manual

Restrictor, Prox Flow (R4)

Wye, patient circuit

Solenoid Valve, Prox Flow (SOL 6)

Sensor, Proximal Flow

Module, Proximal Flow System

Filter, neonatal expiratory

Pressure Sensor, Prox Flow Accumulator (PPROX)

Condensate vial, neonatal expiratory

Humidifier

2-39

Running H/F 1

Page Left Intentionally Blank

2-40

Running H/F 3

3 Installation

3.1

Overview
This chapter contains information for the installation and set up of the Puritan
Bennett 980 Series Ventilator. Before operating the ventilator system, thoroughly read this Operators Manual.
Topics include:

3.2

Safety reminders

Ventilator setup

Battery information

Ventilator operating modes

Preparing the ventilator for use

Tests to perform prior to ventilating a patient

Safety Reminders
WARNING:
Explosion hazard Do not use in the presence of flammable gases. An
oxygen-rich environment accelerates combustibility.
WARNING:
To ensure proper operation and avoid the possibility of physical injury, only
qualified medical personnel should attempt to set up the ventilator and
administer treatment with the ventilator.
WARNING:
To prevent electrostatic discharge (ESD) and potential fire hazard, do not use
antistatic or electrically conductive hoses or tubing in or near the ventilator
breathing system.

3-1

Installation

3.3

3.3.1

WARNING:
Use only gas supply hoses approved by Covidien. Other hoses may be
restrictive and may cause improper ventilator operation.
WARNING:
To avoid possible injury, lock the ventilators casters prior to installing or
removing ventilator components.
Caution:
To ensure optimum performance, Covidien recommends preventive
maintenance be performed by factory-trained Biomedical Engineers per the
schedule specified. Reference Service Preventive Maintenance Frequency, p.
1-2 in the Operators Manual Addendum in this manual.
Note:
US federal law restricts this device to sale by or on the order of a physician.

Product Assembly
How to Assemble Ventilator Components
Ventilator setup should have already been completed by factory-trained service
personnel including successfully passing EST. This manual does not include
ventilator assembly instructions.

3.3.2

Product Power Sources


Using AC Power
The ventilator is normally AC-powered. Reference Connecting the Ventilator to
AC Power, p. 3-5 to connect the ventilator to AC power.

3-2

Operators Manual

Product Connectivity

Using Battery Power

WARNING:
Use only Covidien-branded batteries. Using other manufacturers brands
could result in the batteries operating the ventilator for less than the
specified amount of time or could cause a fire hazard.
WARNING:
One primary battery must be installed at all times in the BDUs primary
battery slot for proper ventilator operation. The ventilator will not complete
the startup process without the primary battery installed. Reference Battery
Compartment Locations, p. 3-21 for identification of battery slots.

The ventilators primary battery must be installed by qualified service personnel


(as it is shipped separately) before patient use. The ventilator will not complete
Power on Self Test (POST) if the battery is not present, and ventilation is prohibited. Ensure the battery is fully charged before placing the ventilator into
service.
The ventilator employs a battery backup system if AC power becomes unavailable or drops below approximately 90 volts. A new, fully charged battery provides at least one hour of power to the ventilator assuming ambient
temperature of 20C (68F) to 25C (77F), PBW = 70 kg, and at factory
default ventilator settings.
The battery back-up system for the ventilator contains one primary battery.
Backup power is supplied to the ventilator in the event of an AC power loss.
One extended battery slot is available for the ventilator. If both primary and
extended ventilator batteries are present, these batteries can power the ventilator for two hours (one hour for the primary battery and one hour for the
extended battery) under the environmental conditions described above. When
using battery power, the ventilator operates from its extended batteries, if
present, first and then switches to the primary batteries. The ventilator primary
and extended batteries are charged whenever the ventilator is plugged into AC
power (the ventilator does not have to be powered up). If the ventilator is operating on battery power, the status display shows which battery is in use and its
charge level, and the remaining time the battery will operate before charging
is required again.

Operators Manual

3-3

Installation

Battery Charging
Batteries requiring charging are charged whenever the ventilator is connected
to AC power, whether operating or not.
The ventilator charges its primary battery first, then its extended battery. The
time required to charge a single battery (either primary or extended) is approximately six hours at room temperature whether the ventilator is turned off (but
connected to AC power) or operating, but charging time can vary based on
temperature or depletion state of the battery. The status display provides the
batterys capacity.
Green LED bars located on the ends of both primary and extended batteries (if
installed) scroll upwards indicating battery charging. A white LED bar represents the battery is in use and a round LED indicator illuminates red if there
is a battery fault. When running on battery power, battery capacity is determined by the number of green LED bars illuminated. Reference Proper Battery
Orientation, p. 3-20 to view the LEDs. Reference p. 3-19 for information on
interpreting the battery capacity. Green LED bars do not scroll if the battery is
not charging or is in use.
If a battery fault occurs, the fault is annunciated, charging of the faulty battery
discontinues, but charging of any other non-faulty battery continues. A faulty
battery will cause annunciation of the error and battery power will not be available for the ventilator.
The ventilator status display indicates the charge level of the installed batteries,
the presence of one or more battery faults, and which battery is being charged.
The ventilator operates no differently when its batteries are charging than it
does when the batteries are fully charged.
The ventilator continues operating as set when the ventilator switches from AC
power to battery power and illuminates an indicator on the status display alerting the operator that the ventilator is now operating on battery power and AC
POWER LOSS alarm annunciates. A medium priority alarm annunciates when
the remaining run-time for the ventilator drops to ten (10) minutes and a high
priority alarm annunciates when the remaining time drops to five (5) minutes.
3.4

Product Placement
The ventilator is positioned standing on its casters next to the patients bedside, as shown below.

3-4

Operators Manual

Product Connectivity

Move the ventilator using the handle encircling the BDU and roll the ventilator
to the desired location.
Figure3-1.Example of Freestanding Ventilator Placement

3.5

3.5.1

Product Connectivity
Connecting the Ventilator to AC Power
Note:
Power outlet access and power cord position Ensure that the power outlet
used for the ventilator is easily accessible; disconnection from the outlet is the only
way to completely remove power from the ventilator.
To connect the power cord to AC power
1.
Plug the ventilator into a properly grounded power outlet rated for at least 15 A.

Operators Manual

3-5

Installation

2.

Verify the connection by checking the AC indicator below the power switch on
the front of the BDU. Reference Ventilator Power Switch and AC Indicator, p. 226 for the power switch and AC indicator locations.

To connect the power cord to the ventilator


1.
Remove the power cord retainer and connect the female end of the power cord
to the ventilators power cord receptacle. Reference Power Cord Retainer on BDU,
p. 3-7.
2.

Replace the power cord retainer.

Use the power cord hook located at the back of the ventilator for power cord
storage.

3-6

WARNING:
For proper ventilator operation, and to avoid the risk of electric shock,
connect the ventilator to a grounded, hospital grade, AC electrical outlet.

Operators Manual

Product Connectivity

Figure3-2.Power Cord Retainer on BDU

3.5.2

Connecting the Gas Supplies


The ventilator can be connected to hospital grade wall or bottled air and
oxygen. Reference Connecting the Ventilator to the Gas Supplies, p. 3-9. Both
air and O2 supply pressure ranges must be between 35 and 87 psig (241.3 kPa
to 599.8 kPa) and the average flow requirement for both gases is 60 L/min at
40.61 psi. The transient will not exceed 200 L/min for three (3) s.

Operators Manual

3-7

Installation

WARNING:
Due to excessive restriction of the Air Liquide, SIS, and Drger hose
assemblies, reduced ventilator performance levels may result when oxygen
or air supply pressures < 50 psi (345 kPa) are employed.

Gas cross flow from one high pressure input port of one type of gas to another
high pressure input port of a different gas will not exceed 100 mL/h under
normal or single fault conditions. If, during a single fault condition, cross flow
exceeds 100 mL/h, an audible alarm annunciates.

WARNING:
Use of only one gas source could lead to loss of ventilation and/or hypoxemia
if that one gas source fails and is not available. Therefore, always connect at
least two gas sources to the ventilator to ensure a constant gas supply is
available to the patient in case one of the gas sources fails. The ventilator has
two connections for gas sources: air inlet, and oxygen inlet. Reference Nontechnical Alarm Summary, p. 6-18 for alarms that occur due to a loss of either
gas supply.
To connect the gas sources
1.
Connect the oxygen hose to the oxygen inlet fitting (item 1) as shown. Ensure use
of a medical grade oxygen source.
2.

3-8

Connect the air hose to the air inlet fitting (item 2) Reference Connecting the Ventilator to the Gas Supplies, p. 3-9.

Operators Manual

Product Connectivity

Figure3-3.Connecting the Ventilator to the Gas Supplies

O2 gas connection

Air gas connection

WARNING:
To prevent a potential fire hazard and possible damage to the ventilator,
ensure the connections to the gas supplies are clean and unlubricated, and
there is no water in the supply gas. If water is suspected, use an external wall
air water trap to prevent damage to the ventilator or its components.

The ventilator system can be purchased with the following gas inlet fittings for
both air and O2: BOC, DISS, female, NIST, Air Liquide, SIS, and Drger.

Operators Manual

3-9

Installation

Reference Covidien Accessories and Options, p. 9-4 for part numbers of gas
hoses. For countries outside the USA, contact your local Covidien representative for the proper gas hoses.
3.5.3

Filter Installation
The ventilator is shipped with internal and external inspiratory filters. Reference
Covidien Accessories and Options, p. 9-4 for the part numbers of expiratory filters. To prevent infection and contamination, both inspiratory and expiratory
filters must be used with the ventilator.

WARNING:
In order to reduce the risk of infection, always use the ventilator with
inspiratory and expiratory bacteria filters.
WARNING:
Do not attempt to use inspiratory or expiratory filters designed for use with
ventilators other than the Puritan Bennett 980 Series Ventilator. Reference
Covidien Accessories and Options, p. 9-4 for relevant part numbers.
WARNING:
Refer to the filters instructions for use for details such as cleaning and
sterilization requirements, filtration efficiency, proper filter usage, and
maximum filter resistance, particularly when using aerosolized medications.
Caution:
Ensure both inspiratory and expiratory filters are properly attached to the
ventilator.
To install the inspiratory filter
1.
Attach the inspiratory filter to the To Patient port.
2.

3-10

Ensure the direction of flow arrow is pointing outward, toward the patient circuits inspiratory limb.

Note:
Refer to the inspiratory filter IFU for information on proper use and handling of
the filter.

Operators Manual

Product Connectivity

WARNING:
Refer to the expiratory filter instructions for use (IFU) for information on
reusable filter cleaning and sterilization and filter efficiency.
Note:
Refer to the expiratory filter IFU for information on proper use and handling of the
filter and for emptying the condensate vial for adult and pediatric patients. Reference
Appendix E for information on emptying the condensate vial when using neonatal
expiratory filters.
WARNING:
Do not re-use disposable inspiratory or expiratory filters, and dispose
according to your institutions policy for discarding contaminated waste.

The condensate vial must be assembled to the reusable expiratory filter prior
to installing the assembly to the ventilator.
To assemble the Adult/Pediatric reusable expiratory filter and condensate vial
1.
Seat the filter to the condensate vial, ensuring alignment of the condensate vials
seal with the mating edge of the expiratory filter.
2.

Twist the condensate vial in a counterclockwise direction until the stops on the vial
and expiratory filter meet.

To install the Adult/Pediatric expiratory filter


1.
If necessary, remove expiratory limb of patient circuit from expiratory filter.
2.

Raise the expiratory filter latch to unlock (item 6).Reference Adult/Pediatric Filter
Installation, p. 3-12. This raises the exhalation valve assembly and allows the filter
door to swing away from the ventilator.

3.

Open the expiratory filter door.

4.

Remove the existing filter.

5.

Insert the new filter by sliding the filter along the tracks in the door. Ensure the
From Patient port aligns with the cutout in the door and points away from the
ventilator.

6.

Close the expiratory filter door.

7.

Lower the expiratory filter latch to secure the filter.

Operators Manual

3-11

Installation

WARNING:
Do not operate the expiratory filter latch during patient ventilation.
Opening the latch during ventilation will result in a patient disconnect
condition and corresponding alarm.
Figure3-4.Adult/Pediatric Filter Installation

Condensate drain port

Condensate drain port cap

Condensate vial

Expiratory filter latch

Expiratory filter

Expiratory filter door

Condensate vial gasket

To install the neonatal expiratory filter adapter door


1.
If necessary, remove expiratory limb of patient circuit from expiratory filter.

3-12

2.

Lift expiratory filter latch. Reference Installing the Neonatal Filter, p. 3-13 (item 3).

3.

Remove existing expiratory filter door by lifting it off of the pivot pins.

Operators Manual

Product Connectivity

Fit neonatal adapter door onto pivot pins.

4.

Figure3-5.Installing the Neonatal Filter

Neonatal expiratory filter

Expiratory filter latch

Neonatal adapter door

Filter door pivot pin

Note:
Unlike the pediatric/adult expiratory filter and condensate vial, the neonatal filter and
condensate vial assembly is provided assembled. The condensate vial is removable for
discarding accumulated liquid, by turning the vial clockwise to remove and
counterclockwise to install.
To install the neonatal expiratory filter assembly
1.
With the door still open, push the neonatal filter assembly straight up into the
adapter.
2.

Close the door.

3.

Lower the expiratory filter latch.

4.

Re-attach expiratory limb of patient circuit to filter.

To use the drain bag


1.
Remove the drain port cap from the expiratory filter condensate vial drain port.
2.

Operators Manual

Attach the drain bag tube to the condensate vials drain port.

3-13

Installation

3.

Hang the drain bag on the holder located on the ventilators accessory rail, as
shown below. Reference Covidien Accessories and Options, p. 9-4 for part
number of drain bag holder.
Figure3-6.Drain Bag

3.5.4

Connecting the Patient Circuit


Reference Connecting the Adult or Pediatric Patient Circuit, p. 3-16 or Reference Connecting the Neonatal Patient Circuit, p. 3-17 to connect the adult,
pediatric, or neonatal patient circuits.

3-14

Operators Manual

Product Connectivity

WARNING:
Use patient circuits of the lowest compliance possible with the ventilator
system to ensure optimal compliance compensation and to avoid reaching the
safety limit of five times set tidal volume or the compliance compensation
limit. Reference the table below for circuit types corresponding with
predicted body weight (PBW).
Table3-1.Patient Types and PBW Values

Circuit Type

PBW in kg (lb)

Allowed but not recommended

Neonatal

0.3 kg to 7.0 kg (0.66 to 15 lb)

Not applicable

Pediatric

7.0 kg to 24 kg (16 lb to 53 lb)

3.5 kg to 6.9 kg and 25 kg to 35 kg


(7.7 lb to 15 lb and (55 lb to 77 lb)

Adult

25kg to 150 kg (55 lb to 331 lb)

7.0 kg to 24 kg
(16 lb to 53 lb)

Note:
Refer to the patient circuits instructions for use (IFU) for information on proper use
and handling and care and maintenance of the circuit.

A list of breathing system components and accessories is provided. Reference


Covidien Accessories and Options, p. 9-4. Use only Covidien- components and
accessories in the patient circuit.
Follow your institutions protocol for safe disposal of the patient circuit.
Follow the patient circuits instructions for use (IFU) for cleaning and disinfection information for reusable circuits.
Orient the patient circuit by hanging the patient circuit on the circuit management supports provided with the flex arm.

Operators Manual

3-15

Installation

Figure3-7.Connecting the Adult or Pediatric Patient Circuit

3-16

Humidifier

From Patient port

Inspiratory limb

Expiratory filter

Circuit wye

To Patient port

Expiratory limb

Inspiratory filter

Condensate vial

Operators Manual

Product Connectivity

Figure3-8.Connecting the Neonatal Patient Circuit

Humidifier

From patient port

Patient circuit inspiratory limb

Neonatal expiratory filter (installed in


adapter door)

Circuit wye

To patient port

Patient circuit expiratory limb

Inspiratory filter

Condensate vial

Operators Manual

3-17

Installation

3.6

3.6.1

3-18

WARNING:
Do not attempt to sterilize single-patient use circuits.

How to Install Accessories


Batteries
WARNING:
Use only Covidien-branded batteries. Using other manufacturers brands or
remanufactured batteries could result in the batteries operating the
ventilator for less than the specified amount of time or could cause a fire
hazard.
WARNING:
To reduce the risk of infection due to cross-contamination, using a damp
cloth, disinfect the batteries with one of the solutions listed before
installation and whenever transferring to or from another ventilator. During
use, clean external surfaces of batteries as necessary. Reference Surface
Cleaning Agents, p. 7-5. Do not spray disinfectant directly onto the battery or
its connector. Reference Ventilator Battery, p. 3-19 (item 1).
Caution:
Ensure the batteries are oriented properly. Reference Proper Battery
Orientation, p. 3-20.

Operators Manual

Product Connectivity

Figure3-9.Ventilator Battery

Battery connector

Primary Batteries
The ventilators primary battery is located in the rearward battery receptacle on
the right side of the BDU. Reference Battery Compartment Locations, p. 3-21.
The primary battery may be hot swapped, that is it can be replaced while
the ventilator is operating.
To install or replace the primary battery in the BDU
1.
With the battery not installed in the ventilator, or if the ventilator is turned off and
not connected to AC power, check the charge level by pressing the charge level
button on the battery and verifying the charge level LEDs illuminate. Reference
Proper Battery Orientation, p. 3-20. for the location of the charge level button.
Five green LED segments illuminate, indicating 90% battery capacity. From
bottom to top, the first LED indicates 10% capacity, the second LED indicates

Operators Manual

3-19

Installation

25% capacity, the third LED indicates 50% capacity, and the fourth LED indicates 75% capacity. An illuminated red LED at the top of the battery indicates
a battery fault. If no LEDs illuminate it means there is < 10% battery capacity
remaining.
2.

If the charge level is sufficient, orient the battery as shown, Reference Proper
Battery Orientation, p. 3-20, face the front of the ventilator and locate the battery
compartments on the right side of the BDU. Reference Battery Compartment
Locations, p. 3-21. The receptacle towards the rear of the ventilator houses the
primary battery while the receptacle towards the front of the ventilator houses the
extended battery.

3.

The primary battery is fastened in place with a thumbscrew (item 3). Loosen the
thumbscrew approximately four to five turns to allow battery installation.

4.

Insert the battery and push into its receptacle all the way until it clicks, indicating
it is latched. The battery will only fit into the slot one way.
Figure3-10.Proper Battery Orientation

3-20

White battery in-use LED

Green charge status LEDs

Red battery fault indicator

Charge level button

Operators Manual

Product Connectivity

Tighten the thumbscrew to secure the battery and prevent the primary battery
from being removed.

5.

Note:
Remove the primary battery by reversing the steps. After loosening the
thumbscrew, slide the battery ejector to the left to eject the battery.
Figure3-11.Battery Compartment Locations

BDU extended battery receptacle and


battery ejector

BDU primary battery thumbscrew

BDU primary battery receptacle and


battery ejector

BDU primary battery (positioned for


installation)

Operators Manual

3-21

Installation

Extended batteries
The extended battery receptacle is located forward of the primary battery. Like
the primary battery, the extended battery may be hot swapped.
To install or remove an extended battery in the BDU
1.
Properly orient the battery as shown above.
2.

3-22

Push the battery into the forward receptacle in the BDU all the way until it clicks,
indicating the battery is latched. Reference Battery Compartment Locations, p. 321.

Note:
Remove the battery by sliding the battery ejector to the left. The battery ejects
itself from its receptacle. There is no thumbscrew for extended batteries.

WARNING:
Even though the Puritan Bennett 980 Ventilator meets the standards
listed in Chapter 11, the internal Lithium-ion battery of the device is
considered to be Dangerous Goods (DG) Class 9 - Miscellaneous, when
transported in commerce. As such, the Puritan Bennett 980 Ventilator and/
or the associated Lithium-ion battery are subject to strict transport
conditions under the Dangerous Goods Regulation for air transport (IATA:
International Air Transport Association), International Maritime
Dangerous Goods code for sea and the European Agreement concerning
the International Carriage of Dangerous Goods by Road (ADR) for Europe.
Private individuals who transport the device are excluded from these
regulations although for air transport some requirements may apply.

WARNING:
To avoid the risk of fire, explosion, electric shock, or burns, do not short
circuit, puncture, crush, heat above 60C, incinerate, disassemble the
battery, or immerse the battery in water.

Note:
Reference Battery Charging, p. 3-4 for battery charging information when
batteries are installed in ventilator.

Operators Manual

Product Connectivity

3.6.2

Battery Testing
To test the batteries
1.
Push the battery charge level button located on the battery. A series of LEDs illuminates, indicating the charge level of the battery. When the bottom LED is illuminated, there is 10% of full battery capacity. The next LED illuminates when
there is 25% capacity. the third lamp illuminates when there is 50% capacity
available. The fourth LED illuminates when there is 75% capacity, and when the
top LED is illuminated, it represents 90% capacity. Reference Proper Battery Orientation, p. 3-20 to view the battery test button and LEDs.

3.6.3

Battery Performance Test Results


Performance testing on a sample of new batteries and batteries charged and
discharged at least 1000 times was completed to demonstrate that the ventilators LOW BATTERY alarms remain effective. Testing demonstrated that the
batteries have a minimum of ten (10) minutes time remaining from the activation of the low battery alarm and a minimum of five (5) minutes time remaining from the critically low battery alarm until ventilator shutdown. Reference
Status Display Indicators and Descriptions, p. 2-30 for images of the Status
Display during low battery and critically low battery conditions.
Performance testing on a sample of new batteries and batteries charged and
discharged at least 1000 times was completed to demonstrate the expected
run time of the ventilator on battery. This testing was performed for both
typical ventilator settings and adult high demand ventilator settings.
The typical ventilator settings used were:

Operators Manual

Ventilator settings

Assist/Control Ventilation with Volume Control (VC) mandatory type

Tidal volume (VT) = 500 mL

Peak flow (VMAX) = 30 L/min

Respiratory rate (f) = 20 1/min

PEEP = 8 cmH2O

Oxygen concentration (FiO2) = 60%

Flow trigger (V-TRIG) = 3 L/min

3-23

Installation

3-24

Approximate respiratory monitored parameters during simulation

Peak pressure (PPEAK) = 27 cmH2O

Plateau pressure (PPL) during manual inspiratory pause = 23 cmH2O

Exhaled tidal volume (VTE) 442 mL (BTPS)

Total respiratory rate (RTOT) = 20 1/min

I:E ratio = 1:2

PEEP = 8 cmH2O / Total PEEP (PEEPTOT) during expiratory pause = 8 cmH2O

Exhaled minute volume (VETOT) = 8.84 L/min

Adult high-demand ventilator settings

Assist/Control Ventilation with Pressure Control (PC) mandatory type

Inspiratory pressure (PI) = 55 cmH2O

Inspiratory time (TI) = 0.55 s

Respiratory rate (f) = 60 1/min

PEEP = 35 cmH2O

Oxygen concentration (FiO2) = 60%

Flow trigger (V-TRIG) = 20 L/min

Approximate respiratory monitored parameters during adult high-demand simulation

Peak pressure (PPEAK) = 90 cmH2O

Plateau pressure (PPL) during manual inspiratory pause = 79 cmH2O

Exhaled tidal volume (VTE) 900 mL (BTPS)

Total respiratory rate (RTOT) = 60 1/min

I:E ratio = 1:1

Operators Manual

Product Connectivity

PEEP = 39 cmH2O / Total PEEP (PEEPTOT) during expiratory pause = 49 cmH2O

Exhaled minute volume (VETOT) = 52 L/min

The run time does not vary significantly between typical and heavy load settings. The ventilator can be expected to run approximately 75 minutes at
typical settings with new batteries. When running batteries nearing end of life
(batteries with 1000 charge/discharge cycles were used for this data) the run
time can be expected to be approximately 55 minutes.
3.6.4

Battery Life
Battery life for both primary and extended batteries is approximately three (3)
years. Actual battery life depends on the history of use and ambient conditions.
As the batteries age with use, the time the ventilator will operate on battery
power from a fully charged battery will decrease. Replace the battery every
three (3) years or sooner if battery operation time is insufficient for your usage.

3.6.5

Battery Disposal
The battery is considered electronic waste and must be disposed of according
to local regulations. Follow local governing ordinances and recycling plans
regarding disposal or recycling of the battery.

3.6.6

Flex Arm
Use the flex arm to support the patient circuit between the patient and the
ventilator. Reference Flex Arm Installation, p. 3-26, which illustrates flex arm
installation into the sockets provided.

Operators Manual

3-25

Installation

Figure3-12.Flex Arm Installation

To attach or remove the flex arm


1.
Locate the threaded inserts in the ventilators handle.

3-26

2.

Fasten the flex arm into one of the inserts.

3.

Hang the patient circuit using the circuit management supports included with the
flex arm.

4.

Remove the flex arm by first removing the patient circuit, then un-fastening the
flex arm from the threaded fastener in the handle.

Operators Manual

Product Connectivity

3.6.7

Humidifier
Use the humidifier to add heat and moisture to the inhaled gas. Connect the
humidifier to a hospital grade electrical outlet. Choose the humidifier (type and
volume appropriate for the patient). The humidifier may be mounted with the
humidifier bracket as shown. Reference Covidien Accessories and Options, p.
9-4 for the part number of the humidifier bracket.

WARNING:
Selection of the incorrect humidifier type and/or volume during SST or during
patient ventilation can affect the accuracy of delivered volume to the patient
by allowing the ventilator to incorrectly calculate the compliance correction
factor used during breath delivery. This can be a problem, as the additional
volume required for circuit compressibility compensation could be incorrectly
calculated, resulting in over- or under-delivery of desired volume.
WARNING:
To ensure proper compliance and resistance calculations, perform SST with
the humidifier and all accessories used for patient ventilation installed in the
ventilator breathing system.
WARNING:
Follow the humidifier manufacturers Instructions for Use (IFU) when using a
humidifier with patient ventilation.
Caution:
Follow humidifier manufacturers instructions for use (IFU) for proper
humidifier operation.
To install the humidifier bracket
1.
Attach humidifier bracket to the ventilators accessory rail by placing the bracket
behind the railing and fastening the bracket clamp to the bracket with four (4)
5/32 inch hex screws, capturing the railing between the bracket and the clamp.
Ensure the humidifier mounting slots are facing outward from the ventilator.

Operators Manual

3-27

Installation

Figure3-13.Bracket Installation on Rail

To install the humidifier


1.
Slide the rear of the humidifier into the corresponding slot on the humidifier
bracket, until it is fully seated. Reference Humidifier Installation to Ventilator, p.
3-29. Some humidifiers slide into the narrow slot in the humidifier bracket, and
some humidifiers use the wide slot.

3-28

Operators Manual

Product Connectivity

Figure3-14.Humidifier Installation to Ventilator

2.

Fill the humidifier with water to the desired fill volume.

3.

Install the chamber to the humidifier, connect the patient circuit, then run SST.

Operators Manual

3-29

Installation

3.7

3.7.1

4.

Plug the humidifier into a grounded, hospital grade electrical outlet.

5.

Turn the humidifier on.

Note:
Complete instructions for the humidifier bracket and humidifier installation are given
in the Puritan Bennett 980 Series VentilatorHumidifier Bracket Installation
Instructions, which includes humidifier bracket part numbers and descriptions.

Ventilator Operating Modes


Normal Mode
Normal mode is the default mode used for patient ventilation. The ventilator
enters Normal mode after it has been turned on and POST completes, the ventilator is set up, and breath delivery parameters have been entered. If the clinician chooses, s/he can select Quick Start which uses default values or
institutionally configured breath delivery settings after PBW has been entered.
Entry into Normal mode is not allowed if a primary battery is not detected in
the ventilator BDU, a major POST fault occurs, or there is an uncorrected major
system fault, or uncorrected Short Self Test (SST) or Extended Self Test
(EST) failures or non-overridden alerts.
During Normal mode, the omni-directional LED on the top of the GUI appears
green in color, in a steadily lit state. If an alarm occurs, the LED flashes in a color
corresponding to the priority of the alarm. Reference Alarm Prioritization, p. 616 for details regarding alarm priority. If another alarm occurs concurrently
with an existing alarm, the LED displays the color corresponding to the highest
priority level. If the alarm de-escalates, the latched area (located on either side
of the alarm LED indicator) of the alarm LED displays the color of the highest
priority alarm while the center of the LED displays the color of the current
alarms priority. For more information on specific alarms, touch the logs icon
in the constant access icons area of the GUI.

3.7.2

Quick Start
Quick Start is an extension of Normal mode, where institutionally configured
default settings are applied after the patients PBW or gender and height are
entered and Quick Start is touched to begin ventilation.

3-30

Operators Manual

Product Connectivity

3.7.3

Stand-By State
Stand-By state can be used when the clinician needs to disconnect the patient
for any reason (prior to a suction procedure, for example). The ventilator enters
Stand-By state if a request is made by the clinician, a patient is disconnected
within a fixed time period determined by the ventilator software, and the clinician confirms the patient has been disconnected intentionally. If a patient
becomes disconnected from the patient circuit after the time period elapses,
an alarm sounds and the patient-disconnect sequence is initiated. In Stand-by
state, gas output is reduced to ten (10) L/min to limit gas consumption and to
allow for detection of patient reconnection, and O2 concentration becomes
100% for adult and pediatric circuit types and 40% for neonatal circuit types.
Stand-by state is available in all ventilation modes except during Inspiratory and
Expiratory BUV, Occlusion Status Cycling (OSC), Safety Valve Open
(SVO), or Ventilator Inoperative (Vent Inop) conditions.

Note:
Do not block patient circuit wye while in Stand-By state. If the wye is blocked, the
ventilator detects a patient connection and will attempt to resume normal ventilation.
To enter Stand-By state
1.
Touch the Menu tab on the left side of the GUI. The menu appears.
2.

Touch Stand-By. A Stand-By state pending dialog appears instructing the clinician
to disconnect the patient circuit. A timer starts which allows 30 s to disconnect
the patient.

3.

Disconnect the patient circuit and confirm the disconnection by touching Confirm.
A timer starts which allows 30 s for confirmation of disconnect.

To exit Stand-by state


1.
Reconnect the patient circuit. The ventilator resumes ventilation at the settings in
use before the disconnection.

The following ventilator settings become active during Stand-by state:

Base flow is set to ten (10) L/min

100% O2 for adult/pediatric patients

40% O2 for neonatal patients

During Stand-by state

Operators Manual

3-31

Installation

The exhalation valve is open.

Current ventilator settings are retained in memory.

Flow sensors are monitored to detect patient reconnection.

Patient-related alarms are temporarily suppressed, as described below.

Ventilator settings can be changed, if desired, and will be applied upon patient
reconnection.

The ventilator displays an indicator that it is in Stand-by State, and a timer indicating the elapsed time the ventilator has been in Stand-by state.

Ventilator background checks continue to be made.

The ventilator automatically exits Stand-by state when patient reconnection is


detected, the clinician completes patient setup (if ventilation was mistakenly
started before setup was complete), or the ventilator power is cycled.
Prior to entering Stand-by state, the ventilator measures pressure and flow in
the patient circuit to determine if a patient is attached. If a patient is detected,
the ventilator continues ventilation as set prior to the request, alerts the operator that Stand-by state is pending, and requests the patient be disconnected.
A countdown timer appears alerting the operator of the time remaining to disconnect the patient. After the patient is disconnected, the ventilator requests
confirmation of the disconnection. When the ventilator enters Stand-by state,
a message appears on the GUI, any active alarms are silenced and reset and
the associated alarm reset entries are logged in the Alarm Event Log. Alarm
detection is suspended, and breath delivery is suspended while a bias flow is
maintained for patient detection. During Stand-by state, the ventilator displays
the elapsed time the patient has been without ventilation. Since the ventilator
maintains a bias flow for patient detection, it resumes ventilation at the previous settings when the patient is reconnected. There is no need to touch Exit
Stand-By. Reconnecting the patient returns the ventilator to normal operation.
During Stand-by state, patient data values are not displayed and the LED
located at the top of the GUI cycles between yellow and green. Entry into and
exit from Stand-by state is recorded in the General Event log.

3-32

Operators Manual

Product Connectivity

3.7.4

Service Mode
WARNING:
Before entering Service Mode, ensure a patient is not connected to the
ventilator. Ventilatory support is not available in Service Mode.

Service Mode is used for Extended Self Test (EST), ventilator calibration, configuration, software upgrades, option installation (all of which must be performed by Covidien factory-trained service personnel), and for making
adjustments to institutional settings. All information stored in the individual
logs is available in Service Mode. Service Mode logs include:

System diagnostic Log

System Comm. Log

EST/SST Diagnostic log

Settings Log

Alarms Log

General Event Log

Service log

Patient data Log

Reference the Puritan Bennett 980 Series Ventilator Service Manual for
details about Service Mode logs.
A patient must not be attached to the ventilator when entering Service Mode.
Specific actions must be performed to enter this mode, prior to POST completion.
To access Service Mode
1.
Remove the ventilator from patient usage.
2.

Turn the ventilators power switch ON.

3.

Press and release the Service Mode button (TEST) at the back of the ventilator,
when the Covidien splash screen appears on the status display after powering on
the ventilator. Reference Service Mode Button (TEST), p. 3-34. Reference Status
Display Indicators and Descriptions, p. 2-30 for an image of the splash screen. The
ventilator prompts to confirm no patient is attached.

Operators Manual

3-33

Installation

Figure3-15.Service Mode Button (TEST)

Service Mode button

4.

Wait to enter Service Mode.

5.

Confirm a patient is not connected to the ventilator by touching the corresponding button. The message SERVICE MODE VENTILATION SUPPORT IS NOT AVAILABLE appears on the graphical user interface.

6.

Perform required service.

7.

Touch Exit or turn off the ventilator to exit Service Mode.

Reference the Puritan Bennett 980 Series Ventilator Service Manual for
information on which keys are disabled during EST.
In addition to allowing SST to be run, Service Mode also allows configuration
of various items. Reference Ventilator Configuration, p. 3-35 for a list of institutionally- and operator-configurable items.

3-34

Operators Manual

Product Connectivity

3.8

Product Configuration
The ventilator is shipped configured with factory defaults for new patient
parameters which can be configured to suit institutional preferences. The
operator may configure any desired parameter as long as this option has not
been locked out and rendered unavailable. When configuring the ventilator, it
displays the parameters associated with the operators last configuration. The
following table lists the factory-configured settings, the institutionally-configurable settings, and the operator-configurable settings.

WARNING:
If the ventilator fleet in your institution uses multiple institutionally
configured presets and/or defaults, there can be risks of inappropriate alarm
settings.
Table3-2.Ventilator Configuration
Feature

Factory Configured

Institutionally Configurable

Operator
Configurable

Vital patient
data banner

Large font
patient data
panel

Waveform
layout

Display
brightness
(Light settings)

Alarm volume

Elevate O2
control

Date/time
format

Default mL/kg
ratio

Operators Manual

Configured
by Circuit
Type

User Lockable

X
Cant be
changed in
Normal mode

3-35

Installation

Table3-2.Ventilator Configuration
Feature

3-36

Factory Configured

Institutionally Configurable

New patient
startup
defaults
(including
PBW, vent
type, mode,
mandatory
type, trigger
type, O2%,
elevate O2)

Opacity

Operator
Configurable

Configured
by Circuit
Type

User Lockable

Operators Manual

Product Connectivity

3.8.1

Preparing the Ventilator for Use

Caution:
Do not lean on the GUI or use it to move the ventilator. Doing so could break
the GUI, its locking mechanism, or tip the ventilator over.

Prior to ventilating a patient, configure the GUI so it is capable of displaying all


the desired parameters, information, and patient data. This eliminates the
necessity for taking the patient off the ventilator, as configuration of many of
the items requires the unit to be in Service Mode.
To perform institutional configuration
1.
Enter Service Mode, and confirm no patient is attached by touching Configuration. Reference Service Mode, p. 3-33 for instructions on entering Service

Mode.
2.

Touch Configuration at the top of the screen in Service Mode. A list of buttons
appears allowing configuration of the corresponding parameters.

3.

Reference the sections below for specific instructions on institutional configuration of each parameter.

To return to factory default configuration


1.
Enter Service mode, and confirm no patient is attached by touching Confirm. Reference Service Mode, p. 3-33, for instructions on entering Service Mode.

3.8.2

2.

Touch Configuration at the top of the screen in Service Mode. A list of buttons
appears allowing configuration of the corresponding parameters.

3.

Select the desired modified setting from the left-hand menu options.

4.

Touch Default.

Configuring the GUI


The display can be configured in various ways.Reference Ventilator Configuration, p. 3-35 for the parameters which are factory configured, institutionallyconfigurable and operator-configurable. Once the factory or institutionally
configurable items have been configured, they remain the default values.
Factory configured values cannot be changed, however, if the parameters
listed in the referenced table are institutionally configured, then those values
remain in memory as default settings. If changes are made to operator-config-

Operators Manual

3-37

Installation

urable parameters, they remain in memory during a ventilator power cycle as


long as the same patient is set up when returned to ventilation. If a new
patient is set up, the factory configured values or institutionally-configured
values (if the parameter has been configured) are used. No alarm settings are
institutionally configurable, which prevents changes to factory default alarm
settings. However, the default mL/kg ratio is institutionally configurable, which
can affect the default alarm setting values. Always review the alarm defaults
prior to beginning ventilation, and set appropriately.
Date and Time Format
The date and time may be configured to the institutions preference. The time
can be specified as 12-hour or 24-hour time in HH:MM:SS format with onehour and one-minute resolutions, respectively. The date formats are:

DD-MMM-YYYY where DD is a two-digit day format, MMM is a three-letter


abbreviation for the month, and YYYY is a four-digit representation of the year or

MM-DD-YYYY where MM is a two digit month format, DD is a two-digit day


format, and YYYY is a four-digit representation of the year

The settable date corresponds to the number of days in the set month and
accounts for leap years.
To institutionally configure the ventilators date and time settings
1.

Enter Service Mode, and confirm no patient is attached by touching Configuration. Reference Service Mode, p. 3-33 for instructions on entering Service

Mode.

3-38

2.

Touch Date and Time.

3.

Touch the button corresponding to 12-hour or 24-hour time.

4.

Touch Hour and turn the knob to enter the correct hour.

5.

Repeat for the minutes, and am or pm.

6.

Touch the button corresponding to the date format desired (DD-MM-YYYY or


MM-DD-YYYY).

7.

Touch Accept to confirm the date and time.

8.

If done configuring parameters, exit Service Mode.

Operators Manual

Product Connectivity

Pressure Units
The ventilators pressure units can be configured for hPa or cmH2O.
To institutionally configure pressure units
1.
Enter Service Mode, and confirm no patient is attached by touching Configuration. Reference Service Mode, p. 3-33 for instructions on entering Service

Mode.
2.

Touch Vent Setup.

3.

Touch the button corresponding to the desired pressure units.

4.

If done configuring parameters, exit Service Mode by touching Exit.

Screen Brightness and Keyboard Backlight (Light Settings)


To institutionally configure screen brightness and keyboard backlight
1.
Enter Service Mode, and confirm no patient is attached by touching Configuration. Reference Service Mode, p. 3-33 for instructions on entering Service

Mode.
2.

Touch Light Settings. Sliders appear to adjust the screen brightness and keyboard
backlight.

3.

Move the sliders to increase or decrease the brightness and backlight levels. Alternatively, turn the knob to increase or decrease the brightness and backlight levels.

4.

Touch Accept to apply the changes, or Cancel to revert to original settings.

5.

If done configuring parameters, exit Service Mode.

To adjust display brightness


1.
Press the display brightness key.
2.

Slide the brightness slider or turn the knob to adjust the brightness level.

3.

Dismiss the slider by touching anywhere on the GUI screen or allow to time out in
five (5) s.

Operators Manual

3-39

Installation

New Patient Setup Defaults


To institutionally configure new patient default settings
4.
Touch the button corresponding to adult, pediatric, or neonatal New Patient
Defaults.
5.

Touch the Vent type, Mode, Mandatory type, and Trigger type buttons corresponding to the desired parameters.

6.

Configure the default PBW and mL/kg ratio, Elevate O2 and O2% by touching its
button and turning the knob.

7.

Repeat for each patient type by selecting the corresponding button.

8.

Touch Accept or Accept ALL when the default configuration is complete.

9.

If done configuring parameters, exit Service Mode.

Elevate O2

Note:
The Elevate O2 control adds a percentage of O2 to the breathing mixture for two
minutes. The additional percentage is shown on the icon in the constant access icon
area. The allowable range is 1% to 100%.
To adjust the amount of elevated O2 delivered for two minutes
1.

In the vent setup dialog in Normal mode, touch the Elevate O2 icon in the constant
access icons area of the GUI screen. The icon glows and a dialog appears with a
countdown timer, Elev O2 button highlighted and ready for changes, and Extend,
Stop, and Close buttons.

2.

Turn the knob to increase or decrease the amount of oxygen by the amount
shown on the button. The allowable range is +1% to +100% oxygen.

3.

Touch Extend to extend the two-minute interval. Touching Extend restarts the
two-minute countdown timer.

4.

Touch Stop to stop additional oxygen from being delivered and dismiss the countdown timer.

The Elevate O2 function follows these rules:

3-40

Operators Manual

Product Connectivity

If the current O2 setting is 80% or above and the selected delivery increment is
> 20%, the ventilator will deliver 100% O2 for two (2) minutes, after which the
oxygen sensor will be calibrated as long as the full two-minute interval elapses
without a change in O2 delivery.

If apnea ventilation occurs during the two-minute interval, the apnea % O2 delivery also increases by the configured amount.

During LOSS OF AIR SUPPLY or LOSS OF O2 SUPPLY alarm conditions, the Elevate
O2 function is canceled if in progress, and is temporarily disabled until the alarm
condition no longer exists.

During Safety PCV, the Elevate O2 control has no effect. During circuit disconnect
and stand-by states (when the ventilator is turned on but not ventilating) the
Elevate O2 function affects the currently delivered oxygen concentration, not the
set oxygen concentration.

Alarm Volume

WARNING:
The audio alarm volume level is adjustable. The operator should set the
volume at a level that allows the operator to distinguish the audio alarm
above background noise levels.
To institutionally configure the alarm volume
1.
Enter Service Mode, and confirm no patient is attached by touching Configuration. Reference Service Mode, p. 3-33 for instructions on entering Service

Mode.
2.

Touch Alarm Volume Defaults. A screen appears allowing configuration of the


alarm volume by circuit type.

3.

Slide the alarm slider for each circuit type (adult, pediatric, or neonatal) or turn the
knob to configure the alarm volume. The volume settings range from 1 (minimum)
to 10 (maximum).

4.

If done configuring the alarm volume, exit Service Mode.

To adjust alarm volume


1.
Set the alarm volume by pressing the alarm volume key, then sliding the alarm
volume slider or turning the knob.The alarm values range from 1 (minimum) to 10
(maximum).

Operators Manual

3-41

Installation

Note:
A sample alarm tone sounds for verification at each volume level change. If
necessary, re-adjust the alarm volume by moving the alarm volume slider to
increase or decrease the volume.

Note:
The alarm volume reverts to the institutionally configured default alarm volume or
factory default if the ventilators power is cycled.
2.

Dismiss the slider by touching anywhere on the GUI screen or allow to time out in
five (5) seconds.

Vital Patient Data


Patient data are displayed in the Vital Patient Data banner. The operator can
configure the banner for displaying the desired patient data. Reference Areas
of the GUI, p. 4-3. A total of 14 values may be configured at one time, with
eight (8) values visible, and six (6) more visible by scrolling the values using the
left- and right- pointing arrows in the patient data area.
Two pages of additional patient data may be viewed by touching or swiping
down on the patient data tab at the top of the GUI. Choose the respective
buttons to view page one or page two. Additional patient data values may not
be changed.
Reference Patient Data Range and Resolution, p. 11-19 for default patient data
values.
To institutionally configure patient data displayed on the GUI
1.
Enter Service mode, and confirm no patient is attached by touching Configuration. Reference Service Mode, p. 3-33 for instructions on entering Service

Mode.

3-42

2.

Touch Patient Data Defaults. Five (5) layout preset buttons appear along with a list
of parameters and descriptions.

3.

Touch a preset button and individually select a parameter from the scrollable list
below to appear in that presets vital patient data banner. Use the right- and leftpointing arrows to configure default values for all available parameters.Additionally, touch the padlock icon above each patient data parameter on the data
banner to allow (unlocked) or restrict (locked) operator configurablity of that
parameter during normal ventilation.

Operators Manual

Product Connectivity

4.

When done configuring the selected preset, touch Accept and select another
preset to configure, if desired.

5.

Touch Defaults to return configuration to factory settings.

6.

If done configuring parameters, exit Service Mode by touching Exit.

To configure the patient data displayed on the GUI


1.
Double-tap a patient data parameter at the top of the GUI screen. A menu of
buttons appears identified with patient data parameters.The parameter at the
location touched will be replaced with the new parameter of choice. To view more
parameters, touch the left- or right- pointing arrows to reveal more parameters.
2.

Touch the button corresponding to the replacement parameter. The existing


parameter is replaced with the new parameter.

3.

Repeat steps 1 and 2 for as many parameters as desired.

Displaying Patient Data With a Larger Font


To improve visibility of patient data, a screen is available that appears with a
larger font. Up to 14 data values may be displayed which include:

Institutional default patient data values (if configured)

Remaining user selected patient data values (up to 14, including waveforms and
loops)

To institutionally configure the large font patient data defaults


1.
Enter Service Mode, and confirm no patient is attached by touching Configuration. Reference Service Mode, p. 3-33 for instructions on entering Service

Mode.
2.

Touch Large Font Patient Data Defaults. Five layout presets appear along with a
list of parameters and descriptions.

3.

Touch a preset button and individually select a parameter for each of the desired
patient data values.

4.

Choose the desired scalar and loop waveforms for the large font patient data display. Waveform thumbnails appear in the three right-most cells of the large font
data panel.

5.

Touch any of the padlock icons along the right-most edge of the selected layout
to prevent operator configurability of the selected row.

Operators Manual

3-43

Installation

6.

Touch Accept or Accept ALL when finished.

7.

If factory defaults are desired for a preset, touch Defaults.

8.

If done configuring parameters, exit Service Mode by touching Exit.

To display the large font patient data panel


1.
Swipe the vital patient data banner tab downward or touch the vital patient data
tab. The additional patient data panel appears.
2.

Swipe the additional patient data banners tab downward or touch the additional
patient data banners tab. Patient data appear in a larger font.

3.

Swipe the large font patient data panel tab upward or touch the tab to return to
the banner to its normal font size.

The large font patient data parameters are configured in the same way as
described in the patient data configuration section above.
Waveforms
Green waveforms denote a mandatory inspiration, yellow waveforms denote
exhalation, and orange waveforms denote a spontaneous inspiration.
The GUI can be configured to display up to three waveforms and two loops
simultaneously in the waveform area. Reference Areas of the GUI, p. 4-3. The
allowable waveforms include flow vs. time, pressure vs. time, volume vs. time,
and PSYNC vs. time (if IE Sync is installed and in IE Sync Monitoring). Reference
Appendix C for more information on IE Sync. Allowable loops include pressure
vs. volume and flow vs. volume. The waveforms display 60 seconds of information and can be shown in a redrawing format, or paused with the ability to
enable a cursor to trace the waveform by turning the knob.
To institutionally configure waveforms and loops
1.
Enter Service Mode, and confirm no patient is attached by touching Configuration. Reference Service Mode, p. 3-33 for instructions on entering Service

Mode.

3-44

2.

Touch Graph Defaults. Five (5) layout presets appear along with a list of parameters and descriptions.

3.

Touch a layout preset button. The parameter(s) button outline glows, signifying
that it can be changed. If more than one parameter can be changed, touch that
parameter to make its outline glow.

Operators Manual

Product Connectivity

4.

Select the parameter from the list whose waveform is desired to appear on the
waveforms screen.

5.

Configure each of the graphic display layouts as described above.

6.

Touch the padlock icon above each graphic layout to prevent operator configuration of the selected layout.

7.

If factory defaults are desired for a preset, touch Defaults.

8.

If done configuring parameters, exit Service Mode by touching Exit.

To configure waveforms and loops


1.
Touch Waveform Layout, located below the displayed waveforms or the vent
setup screen. The icon glows and a menu of various waveform layouts appears.
2.

Touch the desired waveform(s) icon to display. The selected waveform(s) appear
on the GUI screen and the dialog closes.

To change the axis scaling


1.
Touch the desired waveform axis.
2.

Turn the knob to change the value. For each axis, turn the knob to the right to
decrease the values, and turn to the left to increase the values.

To pause waveforms
1.
Touch the pause icon, located below the waveforms area. The icon glows yellow
and allows the breath to complete. A cursor appears and travels along the waveform while turning the knob, displaying the x- and y-axis values.
2.

Touch the pause icon again to re-activate the waveform.

Reference To capture GUI screens, p. 5-2 for information on storing waveforms.


Opacity
To institutionally configure screen opacity
1.
Enter Service Mode, and confirm no patient is attached by touching Configuration. Reference Service Mode, p. 3-33 for instructions on entering Service

Mode.
2.

Touch the Opacity icon.

3.

Turn the knob to increase or decrease the opacity.

Operators Manual

3-45

Installation

4.

Touch the padlock icon at the right side of the screen to allow or prevent operator
adjustment of the screen opacity.

5.

Touch Accept to close the dialog.

To adjust the screen opacity


1.
Touch the opacity control icon. The icon glows when the opacity can be changed.
2.

3.9

Turn the knob to increase or decrease the opacity.

Note:
The opacity icon can be found on the vent setup screen and on any of the respiratory
mechanics maneuvers screens.

Installation Testing
Fully charge the batteries before placing the ventilator into clinical use. Reference Battery Charging, p. 3-4 for information on battery charging. Reference
p. 3-19 for the meaning of battery charge status LEDs and Reference p. 3-20
for the location of battery test switch and status LEDs.
Prior to connecting a patient to the ventilator for the first time, a qualified
service technician must have calibrated the ventilators exhalation valve, flow
sensors, and atmospheric pressure transducer and performed and successfully
passed EST. Reference the Puritan Bennett 980 Ventilator Service Manual for
instructions.
In addition, the clinician must also perform SST.

3.9.1

3-46

SST (Short Self Test)


WARNING:
Always disconnect the patient from the ventilator prior to running SST or EST.
If SST or EST is performed while a patient is connected, patient injury may
occur.
WARNING:
Check for circuit occlusion and/or run SST if increased pressures are observed
during ventilation.

Operators Manual

Product Connectivity

WARNING:
When changing any accessories in the patient circuit or changing the patient
circuit itself, run SST to check for leaks and to ensure the correct circuit
compliance and resistance values are used in ventilator calculations.

When a patient is not attached to the ventilator, run SST to check the patient
circuit for:

Gas leaks

Circuit compliance and resistance calculations

SST is a five-minute test and must be run under any of the following conditions:

Prior to ventilating a new patient

When replacing the patient circuit and expiratory filter

When connecting a different patient circuit to the ventilator

When changing the patient circuit type

When installing a new or sterilized expiratory filter

When changing the humidification device type

When adding or removing accessories to the breathing system such as a humidifier or water trap

No external test equipment is required, and SST requires minimal operator participation.
Humidification type and volume can be adjusted after running SST, however
the ventilator makes assumptions when calculating resistance and compliance
if these changes are made without re-running SST. For optimal breath delivery,
run SST after changing humidification type and humidifier volume.
SST results are recorded in the SST results log, viewable in Service mode and in
Normal mode using the configuration (wrench) icon.
Required Equipment

Proposed patient circuit for patient ventilation

Accessories (water traps, etc.)

Expiratory filter and condensate vial

Operators Manual

3-47

Installation

Humidifier, if applicable

Other necessary items include:

A no. 1 stopper to block the patient airway at the patient wye

Two gas sources (air and oxygen) connected to the ventilator) at a pressure
between 35 psi and 87 psi (241.3 kPa and 599.8 kPa

SST Test Sequence


To run SST
1.
Ensure a patient is NOT connected to the ventilator.

3-48

2.

So that the ventilator does not detect a patient connection, ensure that the
breathing circuit wye is not attached to a test lung or covered in any way that
would cause an increase in pressure at the wye.

3.

Turn the ventilator on using the power switch located at the front of the BDU,
below the status display. The ventilator runs POST when the power switch is
turned on. Ensure the ventilator is operating on full AC power. Otherwise, SST test
failures may result.

4.

Wait at least 15 minutes to allow the ventilator to warm up and stabilize to ensure
accurate results.

5.

At the ventilator startup screen, touch SST or the Configure icon (wrench) displayed in the lower right area of the GUI. The SST history log appears along with
Patient Setup, Run Leak Test, and Run All SST buttons.

6.

Connect the patient circuit, filters, condensate vial, and all accessories to be used
in patient ventilation. Ensure the patient wye is not blocked.

7.

Touch Run All SST to perform all SST tests or touch Run Leak Test to perform the
SST Leak test of the ventilator breathing circuit.

8.

Touch Accept to continue or Cancel to go back to the previous screen.

9.

After accepting, touch the Circuit Type button corresponding to the patient circuit
type used to perform SST and to ventilate the patient (adult, pediatric, or neonatal.

10.

Touch the Humidification Type button corresponding to the humidification type


used for patient ventilation. If no humidifier is used, touch HME. If a humidifier is
used, touch Humidification Volume and turn the knob to enter the volume. Reference Humidifier Volumes, p. 3-50 to determine the correct volume to enter.

Operators Manual

Product Connectivity

11.

Touch Accept to start SST.

12.

Follow the prompts. Certain SST tests require operator intervention, and will
pause indefinitely for a response. Reference Individual SST Results, p. 3-51 for a
summary of the SST test sequence and results.

13.

After each test, the ventilator displays the results. If a particular test fails, the test
result appears on the screen and a choice to repeat the test or perform the next
test is given. When all of the SST tests are complete, the SST status screen displays
the individual test results.

14.

To proceed to patient set up, (if SST did not detect an ALERT or FAILURE) touch
EXIT SST, then touch Accept or cycle the ventilators power.

The following table lists the tests performed during SST.


Table3-3.SST Test Sequence
Test step

Function

SST Flow Sensor Cross Check Test

Tests O2 and Air Flow Sensors

SST Exhalation Valve Performance

Calibrates the exhalation valve and creates a table for use


during calculations

SST Circuit Pressure Test

Exercises delivery PSOL.


Checks inspiratory and expiratory autozero solenoids.
Cross-checks inspiratory and expiratory pressure transducers at
various pressures.

SST Leak Test

Tests ventilator breathing system for leaks

SST Exhalation Filter Test

Checks for exhalation filter occlusion and exhalation compartment occlusion.

SST circuit Resistance Test

Checks for inspiratory and expiratory limb occlusions, and calculates and stores the inspiratory and expiratory limb resistance
parameters.

SST circuit Compliance Test

Calculates the attached patient circuit compliance.

SST Prox (if Proximal Flow Option


is installed)

Verifies functionality of Proximal Flow Subsystem

Note:
The humidifier volume setting entered during SST should always be equal to the
chambers or columns empty compressible volume. Please do not enter either the
containers compressible volume when full or the containers water volume when full.

Operators Manual

3-49

Installation

For humidifier containers not listed, please enter the manufacturers published
compressible volume-empty (column A) value during SST.

Table3-4.Humidifier Volumes
Model

Description

(A)
Published Compressible
Volume-Empty
(mL)

Humidifier
Volume Setting
for SST (mL)

Fisher & Paykel

MR225

Infant/Ped, disposable, Manual Feed

300

300

Fisher & Paykel

MR290

Neo/Adult Disposable, Autofeed

380

380

Fisher & Paykel

MR250

Adult, Disposable,
Manual Feed

480

480

Fisher & Paykel

MR210

Adult, Disposable,
Manual Feed

480

480

Fisher & Paykel

MR370

Adult, Reusable,
Manual Feed

725

725

Fisher & Paykel

MR340E

Infant, Reusable,
Manual Feed

310

310

Teleflex (Concha)

382-60

Standard compliance

296

296

Teleflex (Concha)

382-30

Low compliance

300

300

Teleflex (Concha)

385-40

Minimal compliance

164

164

Manufacturer

SST Results
SST reports results for each individual test.Three status indicators identify the
SST results and actions to take for each.

3-50

Pass The individual SST test has met its requirements.

Alert Alerts occur when the ventilator detects one or more non-critical faults.

Fail The individual SST test did not meet its requirements.

Operators Manual

Product Connectivity

Table3-5.Individual SST Results


Test status

Meaning

Response

PASS

Individual SST test passed

No need to do anything, unless


prompted by the ventilator.

ALERT

The test result is not ideal, but is not


critical.
If SST is in progress, it halts further
testing and prompts for decision.

When the system prompts, touch one


of these buttons:

FAIL

The ventilator has detected a critical


problem and SST cannot complete
until the ventilator passes the failed
test.

Repeat Test

Next Test

Exit SST

Eliminate leaks in the ventilator


breathing system and re-run SST. Otherwise, service the ventilator and rerun SST.

SST Outcomes

WARNING:
Choose to override the ALERT status and authorize ventilation only when
absolutely certain this cannot create a patient hazard or add to risks arising
from other hazards.

When SST completes all of the tests, analyze the results.


Table3-6.Overall SST Outcomes
Final Outcome

Meaning

Response

PASS

All SST tests passed.

Touch Patient Setup to set up the


patient for ventilation:

ALERT

The ventilator detected one or more


faults. Choose to override the ALERT
status and authorize ventilation only
when absolutely certain this cannot
create a patient hazard or add to risks
arising from other hazards.

To override the alert, touch Override


SST, then touch Accept.
To exit SST, touch Exit SST.

FAIL

One or more critical faults were


detected. The ventilator enters the
SVO state and cannot be used for
normal ventilation until SST passes.

Check the patient circuit to determine


the problem or restart SST with a different patient circuit.
Touch Repeat Test, Run All SST, or Exit
SST, then touch Accept.

Operators Manual

3-51

Installation

If touching Override SST, observe the following warning:


A single circuit leak test can be run, but the full suite of SST test must successfully pass before releasing the ventilator for clinical use.
If a complete SST is interrupted and ventilation was allowed before starting
SST, normal ventilation is allowed if

SST did not detect any failures or alerts before the interruption, and

no other errors that would prevent ventilation occurred, and

there were no changes to the circuit type at the start of the interrupted SST.

During SST, the ventilator displays the current SST status, including the test
currently in progress, results of completed tests. Test data are available in
Service Mode where applicable or are displayed on the screen. The ventilator
logs SST results, and that information is available following a power failure.
The alarm silence and alarm reset keys are disabled during SST, as well as the
Manual Inspiration, Inspiratory Pause, and Expiratory Pause keys.
3.9.2

EST (Extended Self Test)


The ventilators Extended Self Test (EST) function is designed to verify the ventilators operational subsystem integrity.
All required software support to perform EST is resident on the ventilator. EST
requires approximately ten (10) minutes to complete.

Note:
SST is not part of the EST test suite. To determine patient circuit resistance and
compliance, run SST.

EST Self Test Prerequisites


Follow all identified guidelines when performing the EST self test. Inspect all
equipment required for any self test to ensure it is not damaged in any way.

3-52

1.

Collect all required equipment prior to performing any self test of the ventilator.
Successful self test is not possible without the use of the listed equipment.

2.

Disconnect the ventilator from the patient.

3.

Fully charge the primary ventilator battery.

Operators Manual

Product Connectivity

4.

Connect the ventilator to AC power using the hospital-grade power cord until
completion of any self test.

5.

Ensure the ventilator is powered down.

6.

Ensure both air and oxygen sources register pressure between 35 and 87 psi (241
to 599 kPa).

To perform Extended Self Test (EST) or to access additional service functions,


the ventilator must be in Service Mode. Reference Service Mode, p. 3-33.

Note:
15. While in the Service Mode, normal ventilation is not allowed.

WARNING:
Always disconnect the ventilator from the patient before running EST.
Running EST while the ventilator is connected to the patient can injure the
patient.
WARNING:
A fault identified during this test indicates the ventilator or an associated
component is defective. Rectify the fault and perform any required repairs
prior to releasing the ventilator for patient use, unless it can be determined
with certainty that the defect cannot create a hazard for the patient, or add
to the risks which may arise from other hazards.

Perform EST during any of the listed conditions.

Prior to initial installation and first time usage of the ventilator

Every six months

Before any preventive maintenance

Following ventilator service or repair

As part of the ventilators routine performance verification

During EST, the ventilator displays the current EST status, including the test
currently in progress, results of completed tests, and measured data (where
applicable). The ventilator logs EST results, and that information is available following a power failure. The ventilator disables several offscreen keys located
on the bezel of the GUI during EST.

Operators Manual

Alarm silence

3-53

Installation

Alarm reset

Manual inspiration

Inspiratory pause

Expiratory pause

Run tests either as a group or as single tests for troubleshooting purposes.


Equipment for EST

3.9.3

Covidien gold standard test circuit

Number one stopper

Air and oxygen sources, both at 35 to 87 psi (241 to 599 kPa).

An adult-sized expiratory filter

Note:
Attempts to run EST with a Neonatal filter can cause some EST tests to fail.

Note:
If using Air Liquide*, Drger*, and SIS air/oxygen hose assemblies, certain EST
tests may fail when using supply pressures less than50 psi (345 kPa) based on
excessive hose restriction.

EST Test Sequence


Note:
If the ventilator has not reached normal operating temperature from recent usage,
allow it to warm up for at least 15 minutes in Service Mode prior to running EST to
ensure accurate testing.
To perform EST
1.
Review and perform all self test prerequisites. Reference EST Self Test Prerequisites, p. 3-52.

3-54

2.

Collect the appropriate equipment. Reference Equipment for EST, p. 3-54.

3.

Access Service Mode. Reference Service Mode, p. 3-33.

Operators Manual

Product Connectivity

4.

Verify all three CALIBRATION tests under the CALIBRATION tab have passed.

5.

Touch the SELF TEST tab from the horizontal banner at the top of the monitoring
screen.

6.

Touch the EST tab from the left-hand menu options.

7.

Touch Run All to run all tests in sequence or select the desired individual test.

8.

Choose one of the available options: touch Accept to continue; touch Cancel to
go back to the previous screen; or touch Stop to cancel EST.

9.

Follow the prompt to remove the inspiratory filter and connect the gold standard
circuit.

10.

Touch Accept.

11.

Follow prompts to complete EST. The EST tests require operator intervention, and
will pause indefinitely for a response. Reference EST Test Sequence, p. 3-54.

12.

At the DISCONNECT O2 prompt, disconnect the high pressure oxygen source.

13.

At the CONNECT AIR AND O2 prompt, connect both high pressure air and oxygen
sources.

14.

Touch Run All or select the desired individual test. After each test, the ventilator
displays the results.

15.

If a particular test fails, either repeat the test or perform the next test.

16.

When all of the EST tests complete, review test results by pressing each individual
test listed on the left side of the GUI.

17.

Touch EXIT EST.

18.

Touch ACCEPT. The ventilator reruns POST and then displays the ventilator startup
screen.

Table3-7.EST Test Sequence


EST Test Step
Zero Offset

Operators Manual

Function
Tests inspiratory and expiratory pressure transducers
and flow sensors at ambient pressure.

Required User
Interaction
Follow prompts

3-55

Installation

Table3-7.EST Test Sequence (Continued)


EST Test Step

3-56

Function

Required User
Interaction

Leak Test

Determines ability of system to hold pressure.

Follow prompts

Mix Leak

Verifies integrity of the mix system.

Follow prompts

Mix PSOL

Verifies mix PSOL function.

None

Mix Accumulator

Verifies mix accumulator pressure sensor and overpressure switch function.

None

Circuit Pressure

Checks inspiratory and expiratory autozero solenoids

Cross-checks safety valve, inspiratory and expiratory pressure transducers at various pressures

Verify the autozero solenoids function

None

Flow Sensor Cross


Check Test

Verifies all flow sensors and PSOLs at specified flow


volumes.

None

Delivery PSOL

Verifies delivery PSOL current function.

None

Exhalation Valve
(EV) Loopback

Verifies exhalation valve.current and loopback


current are within range.

None

Exhalation Valve
(EV) Pressure Accuracy

Verifies current versus pressure values in flash


memory correspond with actual installed exhalation
valve.

None

Exhalation Valve
(EV) Performance

Verifies the exhalation valve operates within specifications of the last exhalation valve calibration.

None

Exhalation Valve
(EV) Velocity Transducer

Verifies the velocity transducer is sending a signal and


the control circuit recognizes it. It does not verify the
quality of the signal.

None

Safety System

Tests safety valve operation.

None

Backup Ventilation

Verifies backup ventilation systems: mix, inspiratory,


and exhalation.

None

Communication

Verifies GUI communication ports function, both


serial and ethernet.

None

Internal Storage

Verifies internal storage device function.

None

LCD Backlight

Verifies GUI LCD backlight intensity function.

None

Operators Manual

Product Connectivity

Table3-7.EST Test Sequence (Continued)


EST Test Step
Status Display

3.9.4

Function

Required User
Interaction

Verifies status display function

Verifies LCD function

Communicates with BD CPU

None

GUI Audio

Tests GUI alarm indicators, cycling through each


alarm status indication.

Follow prompts

BD Audio

Verifies BD audible alarm is functional. Also verifies


power fail capacitor can operate loss-of-power alarm.

Follow prompts

Rotary Knob Test

Verifies knob rotation function.

Follow prompts

Offscreen Key Test

Verifies GUI bezel key function.

Follow prompts

Ventilator Battery

Tests ventilator battery and power distribution.

Follow prompts

EST Test Results


Table3-8.Individual EST Results
Test Status

Meaning

Response

PASS

Individual EST test passed

No need to do anything unless


prompted by the ventilator

ALERT

The test result is not ideal, but is not


critical.
If EST is in progress, it halts further
testing and prompts for decision.

When the system prompts, select:


REPEAT TEST,
NEXT TEST, or
STOP,
then touch ACCEPT.

FAIL

EST not successfully passed.

Select:
REPEAT TEST,
NEXT TEST, or
STOP,
then touch ACCEPT.

NEVER RUN

Test still requires successful PASS.

Run all EST tests.

When EST completes all of the tests, analyze the results.

Operators Manual

3-57

Installation

Table3-9.Overall EST Outcomes


Final Outcome

Meaning

Response

PASS

All EST tests passed

EST successfully completed. Select


other SERVICE MODE functions or
prepare for SST tests prior to returning the ventilator for patient usage.

ALERT

The ventilator detected one or more


faults. The test result is not ideal, but
is not critical.

When the system prompts, select:


REPEAT TEST,
NEXT TEST, or
STOP,
then touch ACCEPT.

FAIL

One or more critical faults were


detected. The ventilatorenters the
SVO state and cannot be used for
normal ventilation until SST passes.
Service is required.

When the system prompts, select:


REPEAT TEST,
NEXT TEST, or
STOP,
then touch ACCEPT.

OVERRIDDEN

ALERT status overridden by user.

Select next desired test.

Touching Override Est results in the following warning:

3.10

WARNING:
Choose to override the ALERT status and authorize ventilation only when
absolutely certain this cannot create a patient hazard or add to risks arising
from other hazards.

Operation Verification
Before ventilating a patient, you must perform SST and alarms tests with
passing results. Reference To run SST, p. 3-48 and Reference Alarm Testing,
p. 6-9 as well.

3-58

Operators Manual

4 Operation

4.1

Overview
This chapter describes Puritan Bennett 980 Series Ventilator operation and
includes the following sections:

4.2

Setting up the ventilator

How to use the ventilator

How to use the ventilators graphical user interface (GUI)

How to set or change main, alarm, or apnea settings

How to test alarms

How to calibrate, enable, or disable the O2 sensor

How to perform inspiratory and expiratory pause maneuvers.

How to use non-invasive ventilation (NIV)

Ventilator Function
Air and oxygen from wall sources or cylinders enter the ventilator and flow
through individual oxygen and air flow sensors. The gases are then mixed in the
mix modules accumulator. A pressure-relief valve in the mix modules accumulator prevents over-pressurization. The mix module also contains an oxygen
sensor which monitors the air-oxygen mixture according to the operator-set
O2% setting.
After the gas mixes, it flows to the inspiratory pneumatic system, where the
breath delivery flow sensor measures the gas flow and controls a PSOL valve
for proper breath delivery tidal volumes and pressures. The inspiratory pneumatic system contains a safety valve to avoid over-pressure conditions before
flowing through bacteria filters to the patient through the inspiratory limb of
the patient circuit. Upon exhalation, gas flows out the patient circuit expiratory
limb, through the expiratory bacteria filter, through the exhalation valve,

4-1

Operation

which includes the exhalation valve flow sensor, and through the exhalation
port.
4.3

Ventilator Setup
WARNING:
To avoid interrupted ventilator operation or possible damage to the
ventilator, always use the ventilator on a level surface in its proper
orientation.
To set up the ventilator
1.
Connect the ventilator to the electrical and gas supplies. Reference Power Cord
Retainer on BDU, p. 3-7 and Reference Connecting the Ventilator to the Gas Supplies, p. 3-9.

4.4

2.

Connect the patient circuit to the ventilator. Reference the figures on p. 3-16 and
p. 3-17 to connect the adult/pediatric or neonatal patient circuits, respectively.

3.

Turn the ventilator ON using the power switch. Reference Ventilator Power Switch
and AC Indicator, p. 2-26.

4.

Before ventilating a patient, run SST to calculate the compliance and resistance
with all items included in the patient circuit. Reference To run SST, p. 3-48.

User Interface Management


The user interface is structured with a GUI and a status display. The GUI provides
access to ventilator controls and patient data. The status display is a small LCD
panel which acts as a back up to the GUI in the event of a GUI failure. Reference
Status Display, p. 2-28 for more information about the status display.
The status display is not interactive.
During normal ventilator operation, the following information appears on the
status display:
Current power state (AC or DC)

Batteries installed / charge status

4-2

Visual indication of audible alarm volume

Circuit pressure graph displaying PPEAK, PEEP, and pressure-related alarm settings

Operators Manual

Ventilator Shutdown

Reference Status Display, p. 2-28 for information about displayed items during
Service mode.

Using the GUI


The GUI is used to interact with the ventilator while it is ventilating a patient
or in any of its operating modes.

Caution:
Do not lean on the GUI or use it to move the ventilator. Doing so could break
the GUI, its locking mechanism, or tip the ventilator over.

The GUI is divided into several areas.


Figure4-1.Areas of the GUI

1.

Prompt area Located beneath the waveforms. Any prompts or messages


display here. Examples include soft and hard bound messages, PAV+ and IE Sync
startup messages, Oxygen sensor calibration in progress and various other informational messages.

Note:
A soft bound is a selected value that exceeds its recommended limit and requires
acknowledgment to continue. Hard bounds have minimum and maximum limits

Operators Manual

4-3

Operation

beyond which values cannot be selected, however if the desired value is equal to
a settings hard bound, then it is allowable.
2.

Menu tab Located on the left side of the GUI screen. Swiping the tab to the
right and touching Setup causes the Vent, Apnea, Alarm, and More Settings tabs
to appear. Touching those tabs opens screens so that changes to ventilator settings, apnea settings and alarm settings can be made.

3.

Waveform area Located in the center of the GUI screen. Shows various
breath waveforms. Reference To configure waveforms and loops, p. 3-45 for
information on how to configure graphics.

4.

Breath Phase Indicator During normal ventilation, the GUI displays a breath
indicator in the upper left corner which shows the type of breath [Assist (A), Control
(C), or Spontaneous (S)] currently being delivered to the patient, and whether it is in
the inspiratory or expiratory phase. The breath indicator is updated at the beginning
of every inspiration, and persists until the next breath type update. During inspiration,
assist (A) and control (C) breath indicators glow green and spontaneous (S) breath indicators glow orange, each appearing in inverse video where the indicator appears black
surrounded by the colored glow. Reference Areas of the GUI, p. 4-3. During the expiratory phase the breath indicators appear as solid colors (green during assist or control
breaths and orange during spontaneous breaths).

5.

Vital Patient data banner Located across the top of the GUI screen. The
patient data banner displays monitored patient data and can be configured to
show desired patient data. Reference Vital Patient Data, p. 3-42 for information
on configuring patient data for display.

6.

Alarm banners Located on the right side of the GUI screen. Indicates to the
operator which alarms are active, and shown in a color corresponding to priority
(high is red and flashing, medium is yellow and flashing, low is yellow and steady).

7.

Constant access icons Located at the lower right of the GUI screen. This area
allows access to home (house), configure (wrench), logs (clipboard), elevate
oxygen percentage (O2), and help (question mark) icon. These icons are always
visible regardless of the function selected on the GUI.

8.

Constant access area Consists of the Current Settings area and the Constant
access icons. This area allows access to any of the patient setup variables shown
in these areas. Touching an icon causes the particular menu for that variable to
appear.

9.

Current settings area Located at the lower center of the GUI screen. The
ventilators current active settings display here. Touching any of the current settings buttons causes a dialog to appear, allowing changes using the knob.

10.

Vent Setup Button Located at the lower left of the GUI screen. Touching this
button allows access to the ventilator setup screen.

Adjusting GUI Viewing Properties

4-4

Operators Manual

Ventilator Shutdown

Screen Opacity
The opacity control enables the operator to adjust the opacity of the displayed
information between 50% and 100%. At 50%, the displayed image is semitransparent, and at 100%, the displayed image is opaque. The opacity value
remains as set if power is cycled and if the same patient is ventilated. If a new
patient is ventilated, the opacity defaults to 85%. Reference To adjust the
screen opacity, p. 3-46 for instructions on adjusting this feature.
Pushpin Feature
The pushpin feature prevents a dialog from closing under certain conditions
when it is pinned. Like the opacity control, the pushpin appears on the settings
screen after a new patient starts ventilation.
Figure4-2.Pushpin Icon

1
1

Pushpin icon unpinned state

2
2

Pushpin icon pinned state

To use the pushpin


1.
When a dialog is open, for example, if Accept or Accept ALL buttons are available,
touch the unpinned pushpin icon to pin the dialog and hold it open.
2.

Touch Close to close the dialog.

Display Brightness
Display brightness can be controlled manually.This feature is institutionally
configurable. Reference Screen Brightness and Keyboard Backlight (Light Settings), p. 3-39. The brightness range is from 1% to 100% with 1% resolution.
The default value is 80%.
To manually adjust display brightness
1.
Press the display brightness key.
2.

Operators Manual

Slide the brightness slider to the right to increase the brightness level or to the left
to decrease the brightness level. Alternatively, turn the knob to increase or
decrease the brightness level. The control disappears from the screen in approximately five (5)seconds.

4-5

Operation

Display Lock
The primary display provides a display lock key to prevent inadvertent changes
to settings. When active, the display lock disables the touch screen, knob, and
off-screen keys (other than the display lock key) and illuminates an LED on the
display bezel. An image of the display lock icon appears transparently over anything displayed on the GUI, should the operator attempt to use the GUI. Any
new alarm condition disables the display lock and enables normal use of the
GUI.
To lock and unlock the display
1.
Press the display lock key on the GUI. The keyboard LED illuminates and a transparent locked icon appears on the screen, indicating display lock. The icon shortly
disappears, but if the operator tries to activate any of the touch screen controls,
the icon re-appears.
2.

To unlock the display, press the display lock key again. The display lock LED turns
off and an unlocked image briefly appears on the screen.

Using Gestures When Operating the GUI


The GUI incorporates a gesture-based interface where features can be actuated with the fingers using different motions. The following table explains gestures used with the GUI.
Table4-1.Gestures and Their Meanings
Gesture
Swipe

4-6

Description

Used for

How to Use

Quickly brush the


screen surface
with the fingertip.

Opening or closing dialogs


or panels that slide in and
out from the screen sides or
top, moving waveform data,
expanding or collapsing
tooltips, scrolling lists, or
alarm banners, maximizing
or minimizing waveforms.

Swipe toward the center of the screen to


open dialogs or panels. Swipe toward the
side of the screen (or upward if viewing the
additional patient data or large font patient
data panels) to close.
To move a paused waveform, swipe in the
desired direction.
Swipe upward anywhere on a waveform to
maximize it, and swipe downward on the
maximized waveform to minimize it.
Swipe a tooltip upward to expand to a long
description and downward to collapse to a
short description. A downward swipe anywhere in the patient data area opens the
additional patient data panel, and another
swipe on the additional patient data tab displays the large font patient data panel.

Operators Manual

Ventilator Shutdown

Table4-1.Gestures and Their Meanings (Continued)


Gesture

Description

Used for

How to Use

Double-tap

Rapidly touch the


screen surface
twice with one
finger.

Maximizing or minimizing
the viewable area of a
dialog, control, or waveform, expanding or collapsing tooltips

Double-tapping maximizes the viewable


waveform area or shows the long description of a tooltip. Double-tapping again minimizes the viewable waveform area or
shows the short description of a tooltip. If
the control is configurable, double tapping
produces the configuration pop-up menu.

Drag

Move the fingertip over the


screen surface
without losing
contact.

Changing x- and y- axis


scales, moving the waveform cursor, moving scrollbars, scrolling lists. Scrolling
speed varies depending
upon how far outside the list
boundary the finger is positioned.

Touch the axis and drag to the right to


increase the waveform x-axis scale, and to
the left to decrease. Touch the axis and
drag upward to increase the y-axis scale and
downward to decrease.
To move the cursor (when the waveform is
paused), touch the cursor and drag it right
or left. The graph responds similarly.
Scroll a list by dragging the scrollbar right or
left or up or down. The list scrolls according
to the direction of the finger movement.
An automatic scrolling feature starts if the
finger is dragged from the inside of a list to
outside its boundary. The farther outside
the boundary the finger is dragged, the
faster the list scrolls.

Touch and
hold

Touch an item
and hold for at
least 0.5 seconds.

Displaying a tooltip dialog


on whatever item is
touched. The tooltip
appears to glow indicating
the touch and hold action.

N/A

Drag and
drop

Touch and drag


an item to
another location
and lift finger to
drop.

Dragging help icon to


describe an onscreen item.

Drag the help icon, located at the lower


right of the GUI screen, to the item in question and drop. If a blue glow appears, a
tooltip is available and appears with information about that item (for example, a
control or symbol).

4.5

Ventilator Operation
Caution:
Do not set containers filled with liquids on the ventilator, as spilling may
occur.

Operators Manual

4-7

Operation

After turning on the ventilator, it will display a Covidien splash screen, and
run Power On Self Test (POST). After the splash screen appears, the ventilator
gives a choice to ventilate the same patient or a new patient, or run SST.
Ventilation parameters are entered via the graphical user interface
(GUI) using the following general steps:

1.

Touch the setting displayed on the GUI.

2.

Turn the knob to the right to increase or to the left to decrease the value.

3.

Touch Accept to apply the setting or Accept ALL to apply several settings at once.

Note:
Quick Start allows for rapid setup and initiation of mechanical ventilation. Review
Quick Start parameters and ensure they are consistent with institutional practice
before using this feature.
WARNING:
Prior to patient ventilation, select the proper tube type and tube ID.
To use Quick Start
1.
Touch New Patient.

4-8

2.

Touch the highlighted PBW button or Gender/Height.

3.

Turn the knob to adjust the patients PBW or gender and height (if gender is
selected, the height selection becomes available).

4.

Touch Quick Start.

5.

Connect the circuit wye adapter to the patient's airway or interface connection.
The patient is ventilated with the institutionally configured or factory configured
QUICK START defaults according to the PBW or gender/height entered, and circuit
type used during SST. There is no prompt to review the settings and the waveforms display appears.

Note:
Connecting the circuit wye adapter to the patient's airway or interface connection
prior to making the ventilation settings causes the ventilator to begin ventilation using
Safety Pressure Control Ventilation (Safety PCV) and annunciate a PROCEDURE ERROR
alarm. As soon as the ventilator receives confirmation of its settings (by touching
Accept or Accept ALL), it transitions out of safety PCV, resets the alarm, and
delivers the chosen settings. Reference Safety PCV Settings, p. 10-72 for a listing of

Operators Manual

Ventilator Shutdown

these settings.
To resume ventilating the same patient
1.
Touch Same Patient on the GUI screen. The previous ventilator settings are displayed on the screen for review prior to applying the settings to the patient.
2.

If the settings are acceptable, touch Accept to confirm. To change any settings,
touch the setting, turn the knob clockwise to increase the value of the setting or
counter-clockwise to decrease the value of the setting, and touch Accept to confirm. To make several settings changes at once, make the desired changes, then
touch Accept ALL to confirm. The appearance of the settings changes from white,
non-italic font showing the current setting to yellow italics (noting the pending
setting). After the settings are accepted, the appearance changes back to white
non-italic font.

3.

Connect the circuit to the patients airway to initiate ventilation.

To ventilate a new patient


1.
Touch New Patient on the GUI screen. The New Patient settings screen appears to
enter the ventilation control parameters. Reference Ventilator Settings Range and
Resolution, p. 11-8 for default ventilator parameter settings.
Figure4-3.New Patient Settings

2.

Operators Manual

Enter the patients PBW or gender and height (if gender is selected, the height
selection becomes available).

4-9

Operation

4.5.1

3.

If the default ventilator settings are appropriate for the patient, touch START to
confirm the settings, otherwise, touch a ventilator setting and turn the knob to
adjust the parameter. Continue this process for all parameters needing adjustment.

4.

Touch Accept or Accept ALL to confirm the change(s).

5.

Connect the circuit to the patients airway to start ventilation.

Ventilator Settings
WARNING:
The ventilator offers a variety of breath delivery options. Throughout the
patient's treatment, the clinician should carefully select the ventilation mode
and settings to use for that patient based on clinical judgment, the condition
and needs of the patient, and the benefits, limitations and characteristics of
the breath delivery options. As the patient's condition changes over time,
periodically assess the chosen modes and settings to determine whether or
not those are best for the patient's current needs.

The following ventilator settings appear at the new patient setup screen:

4-10

Predicted Body Weight (PBW) Adjust the patients PBW, or select the
patients gender and height. Reference Predicted Body Weight (PBW) Calculation,
p. 4-21.

Ventilation type Determines the type of ventilation to be delivered [Invasive or


Non-invasive (NIV)]

Invasive Conventional ventilation using endotracheal (ET) or tracheostomy


(trach) tubes

Non-invasive (NIV) Ventilation using non-vented full-face masks, nasal


masks, infant nasal prongs, or uncuffed ET tubes. Reference Non-invasive
Ventilation (NIV), p. 4-22

Mode Specify the breathing mode (A/C (assist/control), SIMV (synchronous


intermittent mandatory ventilation), SPONT (spontaneous ventilation), BiLevel (if
the BiLevel option is installed), or CPAP. CPAP is only available when circuit type
is neonatal and vent type is NIV. Reference Ventilator Settings Range and Resolution, p. 11-8, Mode setting.

Mandatory type Select PC (pressure control), VC (volume control), or VC+


(volume control plus)

Operators Manual

Ventilator Shutdown

Spontaneous type If SIMV or BiLevel was selected as the Mode, specify PS


(pressure support) or TC (tube compensation. If SPONT was selected as the Mode,
specify PS (Pressure Support), TC (Tube Compensation), VS (Volume Support) or
PAV+ (Proportional Assist Ventilation) (if the PAV+ software option is installed).

Note:
VS, PAV+, and TC are only available during INVASIVE ventilation.

Trigger type Select pressure- triggering (P-TRIG), flow-triggering (V-TRIG), or IE


Sync- triggering (if the IE Sync option is installed, and only in SPONT mode). Reference Appendix C in this manual for detailed information on the IE Sync software
option. Pressure-triggering and IE Sync-triggering are not available when vent
type is NIV. If ventilating a neonatal patient, only flow triggering is available.

Other ways to access the vent setup screen:

Touch Vent Setup at the bottom left of the GUI display

Swipe the menu tab on the left side of the GUI and touch Setup.
Figure4-4.Open Menu Tab

Operators Manual

Setup button

4-11

Operation

Figure4-5.New Patient setup Screen

need new screenshot


To enter settings into the ventilator
1.
Select Vent type, Mode, Mandatory type, Spontaneous type and Trigger type by
touching the corresponding button.
2.

Touch the ventilator setting button needing changes.

3.

Adjust the setting value.

4.

Continue in this manner until all changes are made, then touch Accept or Accept
ALL.

5.

Touch START. Ventilation does not begin until the breathing circuit is connected
to the patients airway. After ventilation begins, waveforms begin plotting on the
displayed waveforms axes. Reference Waveforms, p. 3-44 for information on
setting up the graphics display.

If changes to any settings are required, return to the Vent Setup screen as
described above, or touch a setting icon in the current settings area. Reference
Areas of the GUI, p. 4-3, item 9.

4-12

Operators Manual

Ventilator Shutdown

Note:
A yellow triangle icon appears on tabs and buttons displayed on the GUI containing
unread or un-viewed items. When the item containing the icon is touched, the icon
disappears.
Note:
To make any settings changes after completing patient setup, touch the Vent tab on
the left side of the Setup dialog and make settings changes as described above. The
current setting appears in white font and changes to yellow italics to note the new
value is pending.Touch Accept or Accept ALL to confirm a single change or a batch of
changes. Once the settings are accepted, their appearance changes to white font.
Note:
Selecting QUICK START, Accept, Accept ALL or Start from the Setup dialog
implements all settings in ALL four Setup tabs (Vent Setup, Apnea, Alarms, and More
Settings) and dismisses the Setup dialog.

Tube Compensation
Tube Compensation is a spontaneous breath type selected during ventilator
setup. It allows the ventilator to deliver additional positive pressure to overcome the resistance imposed by the patients artificial airway.
Reference p. 4-12 for more information on setting up the ventilator. Reference
Ventilator Settings Range and Resolution, p. 11-8 for details of specific tube
compensation settings.
To enable TC
1.
Touch the VENT tab on the GUI screen. Reference New Patient setup Screen, p. 412.
2.

Touch SPONT for the mode selection.

3.

Touch TC for Spontaneous type.

4.

Finish setting up the ventilator as described (reference p. 4-12 for information on


entering ventilator settings.

5.

Select the tube type (either endotracheal or tracheostomy) and set the tube ID to
correspond to patient settings.

After making the changes, touch Accept to apply the new settings, or Cancel to
cancel all changes and dismiss the dialog.

Operators Manual

4-13

Operation

Adjust Tube Type, Tube ID, and Humidification

WARNING:
To prevent inappropriate ventilation with TC, select the correct Tube Type (ET
or Tracheostomy) and tube inner diameter (ID) for the patients ventilatory
needs. Inappropriate ventilatory support leading to over-or under-ventilation
could result if an ET tube or trach tube setting larger or smaller than the actual
value is entered.
To select new settings for the tube, follow these steps
1.
Touch Vent Setup on the GUI screen to display the Ventilator setup screen.
2.

Touch Tube Type or Tube ID for the value to be changed.

3.

Turn the knob to change the setting.

4.

Make other tube settings, as necessary.

5.

Touch Accept or Accept ALL to apply the new settings, or Cancel to cancel all
changes and dismiss the dialog.

Note:
The tube type and tube ID indicators flash if TC is a new selection, indicating the
need for entry of the correct tube type and tube ID.
.
To select new settings for the humidifier, follow these steps
1.
From the Ventilator setup screen, touch the More Settings tab. A dialog appears
containing selections for humidifier type and volume.
A Humidifier Volume button appears below the selection only if Non-Heated
Expiratory Tube or Heated Expiratory Tube is selected as the humidifier type.

4-14

2.

Turn the knob to enter a value equal to the dry volume of the humidifier chamber
being used.

3.

Touch Accept or Accept ALL to apply the new settings, or Cancel to cancel all
changes and dismiss the dialog.

Operators Manual

Ventilator Shutdown

4.5.2

Apnea Settings
After making the necessary changes to the ventilator settings touch the Apnea
tab on the left side of the Setup dialog. Although changing the apnea settings
is not required, confirm the default settings are appropriate for the patient.
Apnea ventilation allows pressure control or volume control breath types.
Parameters in pressure-controlled apnea breaths include f, PI, TI O2%, and TA.
Volume controlled apnea breath parameters are f, VT, VMAX, Flow pattern,
O2%, and TA.

Note:
If QUICK START is chosen, the apnea tab on the Vent Setup screen shows a yellow
triangle, indicating the apnea settings have not been reviewed.
Figure4-6.Apnea Setup Screen

need new screenshot

To set apnea parameters


1.
Select the desired apnea breath type (PC or VC).
2.

Enter the desired apnea settings in the same manner as for the ventilator settings.

3.

Touch Accept or Accept ALL to confirm apnea settings.

Operators Manual

4-15

Operation

During apnea pressure ventilation, apnea rise time % is fixed at 50%, and the
constant parameter during a respiratory rate change is TI.
4.5.3

Alarm Settings
After accepting the apnea settings, the display returns once more to show the
waveforms. Return to the Vent Setup dialog and touch the Alarms tab on the
left side of the GUI screen or touch the alarm icon in the constant access icons
area of the GUI screen. The alarms screen appears with the default alarm settings. Reference Alarms Settings Screen, p. 4-17. Review and adjust the alarm
settings appropriately for the patient.

4-16

Note:
If Quick Start is chosen, the alarms tab on the dialog shows a yellow triangle,
indicating the alarm settings have not been reviewed.
Note:
Reference Alarm Settings Range and Resolution, p. 11-17 for new patient default
alarm values. These defaults cannot be changed. The clinician can adjust alarm
settings by following the procedure below. The alarm settings are retained in memory
when the ventilators power is cycled and same patient is selected.Otherwise, current
settings revert to new patient defaults when a new patient is selected.

Operators Manual

Ventilator Shutdown

Figure4-7.Alarms Settings Screen

2
1

To adjust the alarm settings


1.
Touch each alarm setting slider of the alarm(s) to change. Alarm settings are available for PPEAK, fTOT, VE TOT, VTE MAND, VTE SPONT, and VTI parameters.

2.

Turn the knob clockwise to increase the value, or counter-clockwise to decrease


the value.

3.

Continue until all desired alarms are set as necessary.

4.

Touch Accept ALL to confirm the alarm settings.

Note:
There is an additional alarm setting for TC, PAV+, VS, and VC+ breath types: High
inspired tidal volume (2VTI). This alarm condition occurs when the inspired tidal
volume is larger than the setting value. A 1VTI alarm will also cause breath delivery to
transition to the exhalation phase to avoid delivery of excessive inspiratory volumes.
WARNING:
Prior to initiating ventilation and whenever ventilator settings are changed,
ensure the alarm settings are appropriate for the patient.

Operators Manual

4-17

Operation

WARNING:
Setting any alarm limits to OFF or extreme high or low values, can cause the
associated alarm not to activate during ventilation, which reduces its efficacy
for monitoring the patient and alerting the clinician to situations that may
require intervention.

Reference To adjust alarm volume, p. 3-41 to ensure alarm volume is adjusted


properly.

4.5.4

Note:
A sample alarm tone sounds for verification at each volume level change. Re-adjust
the alarm volume by moving the alarm volume slider to increase or decrease the alarm
volume.
Note:
Do not block the patient wye while the ventilator is waiting for a patient connection.
Otherwise the blockage could imitate a patient connection.

Alarm Screen During Operation


During ventilator operation, the alarm screen appears with indicators to let the
operator know the current patient data value for each parameter (item 1), the
parameter alarm settings (items 2 and 3), recent range of patient data values
for the last 200 breaths (item 4). If an alarm occurs, the slider and corresponding limit button show a color matching the alarms priority. Reference Alarm
Screen During Operation, p. 4-19.

4-18

Operators Manual

Ventilator Shutdown

Figure4-8.Alarm Screen During Operation

4.5.5

Pointers show current value of patient


data corresponding to the alarm
parameter

Low alarm setting (in this case


4VTE SPONT)

High alarm setting (in this case


2VTE SPONT)

Range of patient data values for the


particular parameter during the last
200 breaths

Making Ventilator Settings Changes


If, during ventilation, settings changes are necessary that dont involve
changes to PBW, Mode, Breath types, or Trigger types, the current settings
area located at the lower portion of the GUI screen can be used. Reference
Areas of the GUI, p. 4-3 for the location of the current settings area.

Operators Manual

4-19

Operation

To change a ventilator setting using the current settings area


1.
In the current settings area, touch the parameter whose value needs to be
changed. A dialog appears containing buttons for all ventilator settings, with the
selected setting highlighted.
2.

Touch and turn the knob for any other settings that need to be changed.

3.

Touch Accept or Accept ALL.

To change a setting using the Vent Setup button


1.
Touch Vent Setup.
2.

Change the settings as described previously.

3.

Touch Accept or Accept ALL to confirm the changes.

The ventilator settings and the alarm settings chosen remain in memory after
the a power cycle, as long as the same patient is chosen when the ventilator is
set up again. If a new patient is being ventilated, the ventilator and alarm settings revert to their default values. If all power is lost (both AC and battery),
the ventilator and alarm settings in effect prior to the power loss are automatically restored if the power loss duration is five (5)minutes or less. If the power
loss lasts longer than five minutes, ventilation resumes in Safety PCV. Ventilator and alarm settings must be reset for the patient being ventilated. Reference
Safety PCV Settings, p. 10-72 for a list of these settings.
To use the Previous Setup button
1.

To return to the previous settings, touch Previous Setup on the GUI screen. The
ventilator restores the main control and breath settings previously used, as well as
the alarm and apnea settings, and prompts a review by highlighting the previous
values in yellow. The ventilator, alarm, and apnea settings tab text is also shown
in yellow and the tabs show a yellow triangle, indicating there are previous settings that have not been reviewed.

2.

If the settings are acceptable, touch Accept ALL.

The Previous Setup button disappears when the previous settings are confirmed and re-appears when ventilating with new settings.
4.5.6

Constant Timing Variable During Respiratory Rate Changes


A breath timing graph appears at the bottom of the setup screen which illustrates the relationship between inspiratory time, expiratory time, I:E ratio,

4-20

Operators Manual

Ventilator Shutdown

respiratory rate, and the effects on breath timing due to flow pattern, tidal
volume, and VMAX during mandatory PC, VC, BiLevel, or VC+ breaths. With
BiLevel, PC and VC+ breaths, three padlock icons are located underneath
the breath timing graph allowing the operator to select, from left to right, TI,
I:E ratio, or TE as the constant variable during rate changes (or TH, TH:TL ratio,
or TL in BiLevel). If the ventilation mode is SPONT, the padlock icons do not
appear, and the breath timing graph only displays TI for a manual inspiration.
If the mandatory type is VC, the icons do not appear, but the breath timing
graph displays TI, I:E ratio, and TE.
To choose a constant timing variable for rate changes
1.
Touch a padlock icon corresponding to the parameter to make constant during
rate changes (this changes the padlocks appearance from unlocked to locked).
The locked parameter glows in the settings area.

4.6

2.

Turn the knob to adjust the parameters value.

3.

Touch Accept.

Predicted Body Weight (PBW) Calculation


Many default ventilator and alarm settings are based on patient PBW. Either
through the entry of height and gender or directly via setting PBW, the PBW
range spans at least 0.3 kg (0.66 lb) through at least the 155 kg (342 lb) male
and the 150 kg (331 lb) female. Understanding how the ventilator operates at
the very low end of the range of PBW requires awareness that an entry or prediction for PBW drives the value of a delivered volume, which has a lower limit
of 2.0 mL (if using the NeoMode 2.0 option).Data for adult male and female
PBW as a function of height were calculated by applying the equations presented on www.ards.net.
Assume the ventilator (via direct height or PBW entry) registered a PBW of 0.3
kg. If a delivered volume of 4 mL/kg (PBW) was specified, the required volume
would equal only 1.2 mL, which is less than the ventilator minimum of 2.0 mL.
At a desired 4 mL/kg, the infants PBW would need to be at least 0.5 kg or the
desired volume must be reset to greater than 4 mL/kg (PBW). Once the PBW
of the premature infant approaches 1.0 kg (2.2 lb), this restriction disappears.
After entering PBW, review and change all settings as needed.
The correlation function PBW = height was derived from the sources referenced. For subjects whose body weight/height data define the range of PBWs

Operators Manual

4-21

Operation

that include the 20- to 23-week gestational-age neonates and the young male
and female adolescent adults at the foot of the ARDS tables, their PBW values
were taken as the 50th percentile numbers in the Fenton tables and the CDC
and NCHS charts and tables, respectively. Note that the Fenton tables provided
the exclusive information for premature and infant data between 20 weeks
and 50 weeks of fetal and gestational growth.123

4.7

Note:
Any repeated values noted in the tables are the result of decimal rounding.

Non-invasive Ventilation (NIV)


Non-invasive ventilation (NIV) is used when the clinician determines a mask or
other non-invasive patient interface rather than an endotracheal tube would
result in the desired patient outcome.

4.7.1

NIV Intended Use


NIV is intended for use by neonatal, pediatric, and adult patients possessing
adequate neural-ventilatory coupling and stable, sustainable, respiratory drive.

4.7.2

NIV Breathing Interfaces


Covidien has successfully tested the following non-vented interfaces with NIV:

Full-face Mask: Puritan Bennett Benefit Full Face Mask (large, part number
4-005253-00), ResMed Mirage Non-Vented Full Face Mask (medium)

Nasal Mask: ResMed Ultra Mirage Non-vented Mask (medium)

Infant Nasal Prongs: Sherwood Davis & Geck Argyle CPAP Nasal Cannula (small),
Hudson RCI Infant Nasal CPAP System (No. 3)

Uncuffed neonatal ET tube: Mallinckrodt Uncuffed Tracheal Tube, Murphy (3.0


mm)

1. Fenton TR, BMC Pediatrics 2003, 3:13. http://www.biomedcentral.com/1471-2431/3/13.


2. Hamill, PV V. 1977 NCHS growth curves for children birth to 18 years for the United States: National Center for Health Stat (Vital
and Health Statistics: Series 11, Data from the National Health Survey; no. 165) (DHEW publication; (PHS) 78 - 1650). 1977.
3. Kuczmarski RJ, Ogden CL, Guo SS, et al. 2000 CDC growth charts for the United States: Methods and development. National Center for Health Statistics. Vital Health Stat 11(246). 2002.

4-22

Operators Manual

Ventilator Shutdown

4.7.3

WARNING:
Use only non-vented patient interfaces with NIV. Leaks associated with
vented interfaces could result in the ventilators inability to compensate for
those leaks, even if Leak Sync is employed.
WARNING:
Full-face masks used for non-invasive ventilation should provide visibility of
the patient's nose and mouth to reduce the risk of emesis aspiration.
WARNING:
When Using NIV, the patients exhaled tidal volume (VTE) could differ from
the ventilators monitored patient data VTE reading due to leaks around the
interface. To avoid this, ensure Leak Sync is installed.

NIV Setup
NIV can be initiated from either the New Patient Setup screen during Vent
start-up or while the patient is being ventilated invasively. Reference the table
below for using NIV patient setup information.

Table4-2.Setting Up a Patient for NIV


To set up a new patient
1.

Turn the ventilator on.

2.

Select New Patient.

3.

Enter patients PBW or gender and


height.

4.

Touch NIV vent type.

5.

Select mode.

6.

Select mandatory type.

7.

Complete ventilator settings, including


apnea and alarm settings.

Operators Manual

To set up a patient currently being ventilated


1.

Touch or swipe the menu tab on the left side of


the GUI.

2.

Touch Vent Setup.

3.

Perform steps 4 through 7 as if setting the ventilator up for a new patient.

4.

Review the settings, including apnea and alarm


settings and change if necessary.

4-23

Operation

4.7.4

Conversion from INVASIVE to NIV Vent Type


WARNING:
For proper ventilation when changing the Vent Type on the same patient,
review the automatic settings changes described. Adjust appropriately based
on the relevant tables.

Some ventilator settings available during INVASIVE ventilation are not available
during NIV. Reference the following table for automatic settings changes
when changing vent type from INVASIVE to NIV.
Table4-3. INVASIVE to NIV on Same Patient
Current INVASIVE setting

New NIV setting

Breath Mode: BiLevel

Breath mode: A/C

Breath Mode: SIMV or SPONT

High TI SPONT (2TI SPONT) limit setting available

Mandatory Type: VC+

Mandatory type:
Neonatal: PC
Adult/Pediatric: VC

Spontaneous Type: Any type except PS

Spontaneous type: PS

Trigger type: Pressure

Trigger type: Flow


(Flow triggering is the only allowable trigger type
during NIV)

Alarm settings: 4PPEAK (if applicable), 4VE TOT,

Alarm settings: 4PPEAK, 4VE TOT, 4VTE MAND,

4VTE MAND, 4VTE SPONT, INSPIRATION TOO LONG

4VTE SPONT default to NIV new patient values.

(not user-settable)

Reference Alarm Settings Range and Resolution,


p. 11-17. INSPIRATION TOO LONG alarm not
available.

DSENS

DSENS setting defaults to OFF if Leak Sync is


enabled.

Note:
In any delivered spontaneous breath, either INVASIVE or NIV, if Pressure Support is set
to 0 cmH2O, there is always a target inspiratory pressure of 1.5 cmH2O applied.

When in NIV, the Vent Setup buttons appearance changes, letting the operator know the vent type is NIV.

4-24

Operators Manual

Ventilator Shutdown

Figure4-9.Vent Setup Button NIV Indicating NIV vent type

4.7.5

Conversion from NIV to INVASIVE Vent Type


The table below shows automatic settings changes made when changing vent
type from NIV to INVASIVE.
Table4-4.NIV to INVASIVE on Same Patient
Current NIV setting

New INVASIVE setting

Ventilator settings: 2TI SPONT

N/A

Alarm settings:
4PPEAK, 4VE TOT, 4VTE MAND, 4VTE SPONT

Alarm settings: Default to new patient values


dependent upon selected INVASIVE ventilator
settings. Reference Alarm Settings Range and
Resolution, p. 11-17. INSPIRATION TOO LONG
alarm becomes available.

DSENS

DSENS setting defaults to INVASIVE new patient


value. Reference Ventilator Settings Range and
Resolution, p. 11-8.

Operators Manual

4-25

Operation

4.7.6

High Spontaneous Inspiratory Time Limit Setting


NIV includes a setting in SIMV or SPONT modes for High Spontaneous Inspiratory Time limit (2TI SPONT). When a patients inspiratory time reaches or exceeds
the set limit, the ventilator transitions from inspiration to exhalation, and the
1TI SPONT symbol appears at the lower left on the GUI screen, indicating the
ventilator has truncated inspiration (shown below). The 2TI SPONT setting does
not restrict changes to PBW; if the PBW is decreased, 2TI SPONT may decrease
automatically to remain within its allowable limits.
Figure4-10.2TI SPONT Indicator

WARNING:
No audible alarm sounds in conjunction with the visual 2TI SPONT indicator,
nor does the indicator appear in any alarm log or alarm message.

It is possible the target inspiratory pressure may not be reached if the 2TI SPONT
setting is not long enough, or if system leaks are so large as to cause the ventilator to truncate the breath at the maximum allowable2TI SPONT setting.

4-26

Operators Manual

Ventilator Shutdown

4.7.7

Note:
To reduce the potential for not reaching the target pressure, minimize the leaks in the
system and increase the Rise time % and/or decrease the ESENS setting, if appropriate.

NIV Apnea Setup


Set the patients apnea parameters as described. Reference Apnea Settings, p.
4-15. NIV does not change the way apnea parameters are set.

4.7.8

NIV Alarm Settings


The system initially sets most alarm settings based on the patients PBW.
Review all alarm settings, and change as necessary, but startup does not
require confirmation of the settings.
Figure4-11.Default NIV Alarm Settings

Touch the Alarms tab at any time during ventilation to show the current limits
and the monitored patient value shown in white on the indicating arrows for
each alarm. If an alarm is occurring, the indicator LED color changes based on
alarm priority Reference Alarm Prioritization, p. 6-16 for colors and meanings

Operators Manual

4-27

Operation

of alarms and their priorities.

4.8

Note:
The upper and lower limits of an alarm cannot conflict with each other.
Note:
The upper limits for the spontaneous exhaled tidal volume and mandatory exhaled
tidal volume alarms are always the same value. Changing the upper limit of one alarm
automatically changes the upper limit of the other.

Manual Inspiration
A manual inspiration is an operator-initiated mandatory (OIM) inspiration.
When the operator presses the Manual inspiration key, the ventilator delivers
the currently specified apnea breath setting for spontaneous breaths, or the
currently specified mandatory breath setting for mandatory breaths, either
volume- or pressure-based. A volume-based manual inspiration is compliancecompensated. Pressing the Manual inspiration key while in BiLevel mode will
transition from TH to TL or TL to TH depending on when in the breath cycle the
key was pressed.

4.9

Respiratory Mechanics Maneuvers


To access respiratory mechanics maneuvers
1.
Touch or swipe Menu tab on the left side if the GUI.
2.

4-28

Touch RM.

Operators Manual

Ventilator Shutdown

Figure4-12.RM in Menu Tab

3.

Touch the particular tab for the desired maneuver.


Figure4-13.Respiratory Maneuver Tabs
Need new graphic

Operators Manual

4-29

Operation

4.9.1

4.

Follow the prompts on the GUI screen.

5.

Accept or reject the maneuver results. If the result is accepted, its value is saved.

Inspiratory Pause Maneuver


An inspiratory pause maneuver closes the inspiration and exhalation valves and
extends the inspiratory phase of a single, mandatory breath for the purpose of
measuring end inspiratory circuit pressure in order to calculate inspiratory
Plateau Pressure (PPL), lung Static Compliance (CSTAT), and Static Resistance
(RSTAT) of the respiratory system. Pressures on either side of the artificial airway
are allowed to equilibrate, which determines the pressure during a no-flow
state. A request for an inspiratory pause is rejected during apnea ventilation,
safety PCV, OSC, BUV, and in Stand-by state. Inspiratory pauses are allowed in
A/C, SIMV, BiLevel and SPONT modes. If an inspiratory pause maneuver has
already occurred during the breath, a second inspiratory pause maneuver is
not allowed.
Inspiratory pauses can be classified as automatic or manual. The automatic
inspiratory pause lasts at least 0.5 second but no longer than three seconds. A
manual inspiratory pause starts by pressing and holding inspiratory pause key.
The pause lasts for the duration of the key-press (up to seven seconds).
To perform an automatic inspiratory pause
1.
Press and release the inspiratory pause key on the GUI bezel or touch and release
Start if performing an inspiratory pause from the GUI screen as shown above. The
ventilator performs the inspiratory pause maneuver and displays PPL, CSTAT, and
RSTAT along with the date and time.
2.

Touch Accept or Reject to save or dismiss results. If Accept is touched, the results
are displayed.

Cancel an automatic inspiratory pause maneuver by touching Cancel on the


GUI screen.
To perform a manual inspiratory pause
1.
Press and hold the inspiratory pause key on the GUI bezel or touch and hold Start
on the GUI screen if performing an inspiratory pause from the GUI screen as
shown above. The ventilator prompts that the maneuver has started, and to
release to end the maneuver. The ventilator performs the inspiratory pause
maneuver and displays PPL, CSTAT, and RSTAT, along with the date and time.

4-30

Operators Manual

Ventilator Shutdown

2.

Touch Accept or Reject to save or dismiss results. If Accept is touched, the results
are displayed.

Cancel a manual inspiratory pause maneuver by releasing the Inspiratory Pause


key.
4.9.2

Expiratory Pause Maneuver


An expiratory pause extends the expiratory phase of the current breath for the
purpose of measuring end expiratory lung pressure (PEEPTOT) or total PEEP. It
has no effect on the inspiratory phase of a breath, and only one expiratory
pause per breath is allowed. For I:E ratio calculation purposes, the expiratory
pause maneuver is considered part of the exhalation phase.
During an expiratory pause, both inspiratory and exhalation valves are closed,
allowing the pressures on both sides of the artificial airway to equilibrate. This
allows intrinsic PEEP (PEEPI) to be calculated. PEEPI is PEEPTOT minus the set
PEEP level. An expiratory pause can be either automatically or manually administered, and is executed at the next mandatory breath in A/C, SIMV, or BiLevel
modes. In SIMV, the breath cycle in which the pause becomes active (when the
next scheduled ventilator initiated mandatory (VIM) breath occurs) will be
extended by the amount of time the pause is active. For A/C and SIMV, the
expiratory pause maneuver is scheduled for the next end-of-exhalation prior to
a mandatory breath. In BiLevel the expiratory pause maneuver is scheduled for
the next end-of-exhalation prior to a transition from PL to PH. During the expiratory pause maneuver, PEEPI and PEEPTOT equilibration time values are displayed and regularly updated because stabilization of one of these values can
indicate the pause can be ended. During the expiratory pause, the Apnea Interval TA is extended by the amount of time the pause is active. Expiratory pause
requests are ignored if the ventilator is in apnea ventilation, safety PCV,OSC,
BUV, and Stand-by state. Additionally, SEVERE OCCLUSION alarms are suspended during expiratory pause maneuvers. If flow triggering is active, backup
pressure sensitivity (PSENS) detects patient breathing effort.
Maximum duration for a manual expiratory pause is 15 seconds and three (3)
seconds for an automatic expiratory pause.
During a manual or automatic expiratory pause, PEEPI and PEEPTOT appear on
the GUI with the next VIM to allow the clinician to view when these values are
stabilized, indicating the maneuver can be ended.

Operators Manual

4-31

Operation

To perform an automatic expiratory pause


1.
Press and release the expiratory pause key on the GUI or touch and release Start
if performing the expiratory pause from the GUI screen. The ventilator performs
the expiratory pause maneuver and displays a circuit pressure graph, PEEPTOT, and
PEEPI, along with the date and time.
To perform a manual expiratory pause
1.
Press and hold the expiratory pause key on the GUI bezel or touch and hold Start
if performing the expiratory pause from the GUI screen. The ventilator prompts
that the maneuver has started, and to release the button to end the maneuver.
The ventilator performs the expiratory pause maneuver and displays a circuit pressure graph, PEEPTOT, and PEEPI, along with the date and time.
2.

Accept or reject the pause results.

To cancel an expiratory pause maneuver


1.
Touch Cancel on the GUI screen.
4.9.3

Other Respiratory Maneuvers


To perform other respiratory maneuvers, touch the corresponding tab on the
desired maneuver, and follow the prompts on the GUI screen.

4.10

Oxygen Sensor Function


The ventilator's oxygen sensor monitors O2%. This cell is mounted in the inspiratory module in the BDU and monitors the percentage of oxygen in the mixed
gas delivered to the breathing circuit (it may not reflect the actual oxygen concentration in the gas the patient inspires).
Reference the Puritan Bennett 980 Series Ventilator Service Manual for
instructions on replacing the O2 sensor.
New patient default O2% settings are as follows:

4-32

O2 sensor enabled

Neonatal 40% O2

Pediatric/adult 100% O2

Operators Manual

Ventilator Shutdown

Note:
The oxygen sensor has three states: Enabled, Disabled, and Calibrate. The oxygen
sensor is enabled at ventilator startup regardless if New Patient or Same Patient setup
is selected.
To enable, or disable the O2 sensor
1.

Touch Vent Setup.

2.

Touch the More Settings tab. The more settings screen appears.
Figure4-14.More Settings Screen with O2 Sensor Enabled

4.10.1

3.

Touch the button corresponding to the desired O2 sensor function (Enable or Disable).

4.

Touch Accept.

Oxygen Sensor Life


The O2% setting can range from room air (21% O2) up to a maximum of
100% oxygen. The sensor reacts with oxygen to produce a voltage proportional to the partial pressure of the mixed gas. Since ambient atmosphere contains
approximately 21% oxygen, the sensor constantly reacts with oxygen and

Operators Manual

4-33

Operation

always produces a voltage. The useful life of the cell can also be shortened by
exposure to elevated temperatures and pressures. During normal use in the
ICU, the oxygen sensor lasts for approximately one year the interval for
routine preventive maintenance.
Because the oxygen sensor constantly reacts with oxygen, it requires periodic
calibration to prevent inaccurate O2% alarm annunciation. Once a calibrated
oxygen sensor and the ventilator reach a steady-state operating temperature,
the monitored O2% will be within three percentage points of the actual value
for at least 24 hours. To ensure the oxygen sensor remains calibrated, recalibrate the oxygen sensor at least once every 24 hours.
Typically, the clinician uses an O2 analyzer in conjunction with the information
given by the ventilator. If a NO O2 SUPPLY alarm occurs, compare the O2 analyzer reading with the ventilators O2 reading for troubleshooting purposes.
The ventilator automatically switches to delivering air, only (21% oxygen).
4.10.2

Oxygen Sensor Calibration


The oxygen sensor should be calibrated every 24 hours and before use. The calibration function provides a single-point O2 sensor calibration.
To calibrate the O2 sensor

4.10.3

1.

Touch Vent Setup.

2.

Touch More Settings.

3.

Touch the Calibrate for the O2 sensor. The oxygen sensor calibrates within two
minutes. Reference More Settings Screen with O2 Sensor Enabled, p. 4-33.

Oxygen sensor calibration testing


To test the O2 sensor calibration

4-34

1.

Connect the ventilators oxygen hose to a known 100% O2 source (for example,
a medical-grade oxygen cylinder).

2.

Calibrate the oxygen sensor as described above.

Operators Manual

Ventilator Shutdown

4.11

3.

Connect the ventilator oxygen hose to another known 100% O2 source (for
example, a second medical-grade oxygen cylinder).

4.

Set O2% to each of the following values, and allow one minute after each for the
monitored value to stabilize: 21%, 40%, 90%

5.

Watch the GUI screen to ensure the value for O2 (delivered O2%) is within 3% of
each setting within one minute of selecting each setting.

Ventilator Protection Strategies


The ventilator incorporates a number of strategies to support patient safety.
These include Power On Self-Test (POST), SST and a new strategy called Ventilation Assurance which provides alternate means of ventilation in the case
of certain serious faults in the breath delivery system. The descriptions below
detail the system response to potential failures.

4.11.1

Power on Self Test (POST)


The first strategy is to detect potential problems before the ventilator is placed
on a patient. POST checks the integrity of the ventilator's electronics and prevents ventilation if a critical fault is found. (Reference the Puritan Bennett
980 Series Ventilator Service Manual for a complete description of POST). POST
may detect major or minor system faults which manifest themselves as Device
Alerts. Reference DEVICE ALERT Alarm, p. 6-36 for more information.

4.11.2

Technical Fault
A technical fault occurs if a POST or background test has failed. Reference
Power On Self Test (POST), p. 10-76. Based on the test that failed, the ventilator will either ventilate with current settings, ventilate with modified settings,
or enter the Vent Inop state. A technical fault cannot be cleared by pressing
the alarm reset key. It can only be cleared by correcting the fault that caused
it or if alarm reset criteria have been met.

4.11.3

SST
In addition to characterizing the ventilator breathing circuit, SST performs basic
checks on the ventilator's pneumatic system including the breath delivery

Operators Manual

4-35

Operation

PSOL, the Flow Sensors and the Exhalation Valve. Faults detected during SST
must be corrected before ventilation can be started.
4.11.4

Procedure Error
A procedure error occurs when the ventilator senses a patient connection
before ventilator setup is complete. The ventilator provides ventilatory support
using default Safety Pressure Controlled Ventilation (Safety PCV) settings. Reference Safety PCV Settings, p. 10-72

4.11.5

Ventilation Assurance
During ventilation, the ventilator performs frequent background checks of its
breath delivery sub-system (Reference Safety Net, p. 10-71). In the event that
certain critical components in the pneumatics fail, Ventilation Assurance provides for continued ventilatory support using one of three Backup Ventilation
(BUV) strategies, bypassing the fault to maintain the highest degree of ventilation that can be safely delivered (Reference Background Diagnostic System, p.
10-74 for a full description of the Backup Ventilation strategies).

4.11.6

Note:
Do not confuse BUV with Safety PCV, which occurs when a patient is connected
before ventilator setup is complete, or with Apnea ventilation, which occurs in
response to patient apnea.

Safety Valve Open (SVO)


In the event of a serious fault occurring that cannot be safely bypassed, the
ventilator, as a last resort, reverts to a Safe State. In Safe State, the ventilator
opens the Safety Valve and the Exhalation Valve, allowing the patient to
breathe room air (if able to do so), provided the patient circuit is not occluded,
and the inspiratory PSOL valve is closed. During SVO, the patient (if connected)
can breathe room air through the safety valve after it releases pressure in the
patient circuit. The patient exhales through the exhalation valve with minimal
resistance and the exhalation valve also acts like a check valve, limiting gas
from being drawn in through the expiratory filter or expiratory limb of the circuit. SVO conditions are logged into the event and alarm logs as are the events
leading to the SVO condition. If the condition causing SVO clears, the ventilator clears the SVO state. Patient data do not display on the GUI, but graphics
are still plotted. During SVO, the ventilator ignores circuit occlusions and dis-

4-36

Operators Manual

Ventilator Shutdown

connects. If the condition causing SVO can only be corrected by servicing the
ventilator, the SVO alarm cannot be reset by pressing the alarm reset key.
4.11.7

Ventilator Inoperative (Vent Inop)


Vent Inop occurs when the ventilator detects a catastrophic error which prevents all other safety states from operating. Vent Inop limits pressure to the
patient as the ventilator enters the SVO state, disables (closes) the gas mixing
PSOL valves, and purges the gas mixing system accumulator. The safety valve
is opened and a Vent Inop indicator illuminates and a high priority alarm
annunciates from the primary alarm, and the secondary alarm (continuous
tone) is activated. The ventilator can only exit the Vent Inop state by power
cycling and successfully passing EST. The Vent Inop alarm cannot be reset with
the alarm reset key. All detection and annunciation of patient data alarm conditions is suspended.
During a Vent Inop condition, the inspiratory and expiratory pressure drop
measured at the patient wye does not exceed 6.0 cmH2O at 30 L/min.

4.12

Ventilator Shutdown
When the ventilator power switch is turned OFF, the ventilator executes an
orderly shutdown routine, saving patient data before removing power. If the
ventilator detects a patient connected when the power switch is turned OFF, a
high priority alarm is annunciated and a banner on the display requires the
operator to confirm that a power down was requested. Only after the operator
confirms will the ventilator execute the shutdown command.
All logs are retained in the ventilators memory upon ventilator shutdown.
When the logs reach the maximum number of entries, the oldest values are
overwritten with new values. Reference Ventilator Logs, p. 8-2 for information
on ventilator logs.

Operators Manual

4-37

Operation

Page Left Intentionally Blank

4-38

Operators Manual

5 Product Data Output

5.1

Overview
This chapter describes the features of the Puritan Bennett 980 Series Ventilator designed to provide output to the clinician. This includes language,
methods of displaying and transferring data, types of displayed data, and types
of external device ports. Connectivity to an external patient monitoring system
is also included.

5.2

Language
The language used on the ventilator is configured at the factory.

5.3

Data Display
Displayed data are updated in real-time. The practitioner can display up to 60
seconds of waveform data and pause and capture up to two loops using the
screen capture function. The operator can pause the displays and when the
displays are paused, a cursor appears with the relevant numeric values for the
intersecting points of the cursor and waveform or loop. The scalar waveform
contains a single value, but loops contain both x- and y-axis data. The operator
can move the cursor along the waveform or loop using the knob, and read the
corresponding data. Reference Waveforms, p. 3-44 for details regarding configuring and displaying waveforms.

5.4

Data Transfer
Data from the ventilator can be accessed via USB or RS-232 connectors. The
following data are available for downloading via connection to a remote device
or flash drive:

Waveform images (screen capture function): USB port

5-1

Product Data Output

5.4.1

Waveform data: RS-232 port, USB port with USB to serial conversion capability
(per Comm port configuration)

Results from DCI commands: RS-232 port, USB port with USB to serial conversion
capability (per Comm port configuration)

GUI Screen Capture


Caution:
The USB interface should be used for saving screen captures and interfacing
with an external patient monitor. It should not be used to provide power to
other types of devices containing a USB interface.

Covidien recommends a minimum 128 MB flash drive storage device formatted in the 32-bit file format.
To capture GUI screens
1.
Navigate to the desired screen from which you wish to capture an image (for
example, the waveforms screen). There is no need to pause the waveform before
performing the screen capture.
2.

Touch the screen capture icon in the constant access icons area of the GUI screen.
If desired, navigate to another screen and repeat steps 1 and 2 for up to ten (10)
images. If another image is captured, increasing the queue to eleven images, the
newest image overwrites the oldest image so there are always only ten images
available.

Note:
If the camera icon appears dim, it means that the screen capture function is
currently processing images and is unavailable. When processing is finished, the
camera icon is no longer dim and the screen capture function is available.
To transfer the captured images to a USB storage device
1.
Swipe the Menu on the left side of the GUI. Reference Open Menu Tab, p. 4-11.

5-2

2.

Touch Screen Capture. A list of screen captures appears, identified by time and
date. A slider also appears if more images than shown are present.

3.

Insert a passive USB storage device (flash drive) into one of the USB ports at the
rear of the ventilator. Reference Port Locations, p. 5-18. If more than one USB
storage device is installed in the ventilator, touch the button of the destination
USB device where the image will be copied.

Operators Manual

Data Transfer

5.4.2

4.

In the list of images, touch the image name.

5.

Touch Copy. The image is stored on the destination USB storage device.

6.

Alternatively, touch Select All, and all images in the list are stored on the USB
device and which can then be viewed and printed from a personal computer.

Note:
The file format of screen captures is .PNG.

Communication Setup
To specify the communication configuration for the ventilator
1.
Touch the Configure icon in the constant access icons area of the GUI. A menu
appears with several tabs.
2.

Touch Comm Setup tab. The Comm Setup screen appears allowing three (3) ports
that can be configured. These ports can be designated as DCI, Philips, Spacelabs,
or Waveforms.

Note:
Waveforms can be selected on any port, but only on one port at a time.

Operators Manual

5-3

Product Data Output

Figure5-1.Comm Setup Screen

5.4.3

Comm Port Configuration


Configuring the Comm port allows the ventilator to communicate with devices
listed in the Comm Setup screen, or to capture waveform data (in ASCII
format) from the ventilator.
To configure Comm ports
1.
Touch COM1, COM2, or COM3.
2.

Turn the knob indicating the desired device configuration.

3.

Select the desired baud rate. If waveforms was selected, the baud rate automatically becomes configured to 38400.

4.

Select 7 or 8 data bits.

5.

Select parity of even, odd, or none if data bits = 8.

Connect the device to the previously configured port. Reference Port Locations, p. 5-18 for a description and the locations of the Comm ports.

5-4

Operators Manual

Data Transfer

Note:
When a USB port is configured as a Comm port, it is necessary to use a USB-to-serial
adapter cable. This adapter must be based on the chipset manufactured by Prolific.
For further information, contact your Covidien representative.

Selecting waveforms when configuring a Comm port allows the ventilator to


continuously transmit pressure, flow, and sequence numbers in ASCII format
from the selected serial port, at a baud rate of 38400 bits/s, and the operatorselected stop bits, and parity. A sample of pressure and flow readings is taken
every 20 ms. This sample of readings is transmitted on the selected serial port
at the end of each breath at breath rates of 10/min and higher. For longer
duration breaths, at least the first eight seconds of the breath is transmitted.
The format of the data is as follows: The beginning of inspiration is indicated
by BS, S:nnn,<LF> where 'BS identifies the Breath Start, S:nnn is a
sequence number incremented at every breath, and <LF> is a line feed character. The fff, and ppp fields show the breath flow and pressure data. The end
of exhalation is indicated by: BE<LF> where BE indicates Breath End, and
<LF> is a line feed character.
5.4.4

Serial Commands
The ventilator system offers commands that allow communication to and from
the ventilator using a Comm port. Commands to the ventilator from a remote
device include:

5.4.5

RSET: Reference RSET Command, p. 5-5

SNDA: Reference SNDA Command, p. 5-6

SNDF: Reference SNDF Command, p. 5-10

Note:
The ventilator responds only if it receives a carriage return <CR> after the command
string.

RSET Command
The RSET command clears data from the ventilator receive buffer. The ventilator does not send a response to the host system. Enter the RSET command
exactly as shown:

Operators Manual

5-5

Product Data Output

RSET<CR>
5.4.6

SNDA Command
The SNDA command instructs the ventilator to send information on ventilator
settings and monitored patient data to the host system. Enter the SNDA
command exactly as shown:
SNDA<CR>
When the ventilator receives the command SNDA<CR>, it responds with the
code MISCA, followed by ventilator settings and monitored patient data information.
The MISCA response follows this format:
MISCA

706

97

<STX>

FIELD 5, FIELD 101,

<ETX>

<CR>

Response code to SNDA command

Data field, left-justified and padded


with spaces

Number of bytes between <STX> and


<CR>

End of transmission (03 hex)

Number of data fields between <STX>


and <ETX>

Terminating carriage return

Start of transmission (02 hex)

Fields not available are marked as Not used. Underscores represent one or
more spaces that pad each character string.
The table below lists MISCA responses to SNDA commands.
Table5-1.MISCA Response
Component

5-6

Description

MISCA

Response to SNDA command (5 characters)

706

The number of bytes between <STX> and <CR> (3 characters)

97

The number of fields between <STX> and <ETX> (2 characters

<STX>

Start of transmission character (02 hex)

Operators Manual

Data Transfer

Table5-1.MISCA Response (Continued)


Component

Operators Manual

Description

Field 5

Ventilator time (HH:MM_) (6 characters)

Field 6

Ventilator ID to allow external hosts to uniquely identify each Puritan Bennett


980 Ventilator (18 characters)

Field 7

Room number (6 characters)

Field 8

Date (MMM_DD_YYYY_) (12 characters)

Field 9

Mode (CMV___, SIMV__, CPAP__ or BILEVL) (CMV = A/C) setting (6 characters)

Field 10

Respiratory rate setting in breaths per minute (6 characters)

Field 11

Tidal volume setting in liters (6 characters)

Field 12

Peak flow setting in liters per minute (6 characters)

Field 13

O2% setting (6 characters)

Field 14

Pressure sensitivity setting in cmH2O (6 characters)

Field 15

PEEP or PL (in BiLevel) setting in cmH2O (6 characters)

Field 16

Plateau time in seconds (6 characters)

Field 17 - 20

Not used (6 characters)

Field 21

Apnea interval in seconds (6 characters)

Field 22

Apnea tidal volume setting in liters (6 characters)

Field 23

Apnea respiratory rate setting in breaths per minute (6 characters)

Field 24

Apnea peak flow setting in liters per minute (6 characters)

Field 25

Apnea O2% setting (6 characters)

Field 26

Pressure support setting in cmH2O (6 characters)

Field 27

Flow pattern setting (SQUARE or RAMP__) (6 characters)

Field 28 - 29

Not used (6 characters)

Field 30

Elevate O2 state (ON____ or OFF___) (6 characters)

Field31 - 33

Not used (6 characters)

Field 34

Total respiratory rate in breaths per minute (6 characters)

Field 35

Exhaled tidal volume in liters (6 characters)

Field 36

Exhaled minute volume in liters (6 characters)

Field 37

Spontaneous minute volume in liters (6 characters)

5-7

Product Data Output

Table5-1.MISCA Response (Continued)


Component

5-8

Description

Field 38

Maximum circuit pressure in cmH2O (6 characters)

Field 39

Mean airway pressure in cmH2O (6 characters)

Field 40

End inspiratory pressure in cmH2O (6 characters)

Field 41

Expiratory component of monitored value of I:E ratio, assuming inspiratory component of 1 (6 characters)

Field 42

High circuit pressure limit in cmH2O (6 characters)

Field 43 - 44

Not used (6 characters)

Field 45

Low exhaled tidal volume limit in liters (6 characters)

Field 46

Low exhaled minute volume limit in liters (6 characters)

Field 47

High respiratory rate limit in breaths per minute (6 characters)

Field 48

High circuit pressure alarm status (NORMAL, ALARM_, or RESET_) (6 characters)

Field 49 - 50

Not used (6 characters)

Field 51

Low exhaled tidal volume (mandatory or spontaneous) alarm status (NORMAL,


ALARM_, or RESET_) (6 characters)

Field 52

Low exhaled minute volume alarm status (NORMAL, ALARM_, or RESET_)


(6 characters)

Field 53

High respiratory rate alarm status (NORMAL, ALARM_, or RESET_) (6 characters)

Field 54

No O2 supply alarm status (NORMAL, ALARM_, or RESET_) (6 characters)

Field 55

No air supply alarm status (NORMAL, ALARM_, or RESET_) (6 characters)

Field 56

Not used (6 characters)

Field 57

Apnea alarm status (NORMAL, ALARM_, or RESET_) (6 characters)

Field 58 - 59

Not used (6 characters)

Field 60

Ventilator time (HH:MM_) (6 characters)

Field 61

Not used (6 characters)

Field 62

Date (MMM_DD_YYYY_) (12 characters)

Field 63

Static compliance (CSTAT) from inspiratory pause maneuver in mL/cmH2O (6 characters)

Field 64

Static resistance (RSTAT) from inspiratory pause maneuver in cmH2O/L/s


(6 characters)

Operators Manual

Data Transfer

Table5-1.MISCA Response (Continued)


Component

Operators Manual

Description

Field 65

Dynamic compliance (CDYN) in mL/cmH2O (6 characters)

Field 66

Dynamic resistance (RDYN) in cmH2O/L/s (6 characters)

Field 67

Negative inspiratory force (NIF) in cmH2O (6 characters)

Field 68

Vital capacity (VC) in L (6 characters)

Field 69

Peak spontaneous flow (PSF) in L/min (6 characters)

Field 70

Ventilator-set base flow in liters per minute


(6 characters)

Field 71

Flow sensitivity setting in L/min (6 characters)

Field 72 - 83

Not used (6 characters)

Field 84

End inspiratory pressure in cmH2O (6 characters)

Field 85

Inspiratory pressure or PH setting in cmH2O (6 characters)

Field 86

Inspiratory time or TH setting in seconds (6 characters)

Field 87

Apnea interval setting in seconds (6 characters)

Field 88

Apnea inspiratory pressure setting in cmH2O (6 characters)

Field 89

Apnea respiratory rate setting in breaths per minute (6 characters)

Field 90

Apnea inspiratory time setting in seconds (6 characters)

Field 91

Apnea O2% setting (6 characters)

Field 92

Apnea high circuit pressure limit in cmH2O (6 characters)

Field 93

Alarm silence state (ON____ or OFF___) (6 characters)

Field 94

Apnea alarm status (NORMAL, ALARM_ or RESET_) (6 characters)

Field 95

Severe Occlusion/Disconnect alarm status (NORMAL, ALARM_ or RESET_)


(6 characters)

Field 96

Inspiratory component of I:E ratio or High component of H:L (BiLevel) setting


(6 characters)

Field 97

Expiratory component of I:E ratio setting or Low component of H:L (BiLevel)


(6 characters)

Field 98

Inspiratory component of apnea I:E ratio setting (6 characters)

Field 99

Expiratory component of apnea I:E ratio setting (6 characters)

5-9

Product Data Output

Table5-1.MISCA Response (Continued)


Component

5.4.7

Description

Field 100

Constant during rate setting change for pressure control mandatory breaths
(I-TIME or I/E___ or______) (6 characters) (where ______ represents TE or PCV not
active)

Field 101

Monitored value of I:E ratio (6 characters)

<ETX>

End of transmission character (03 hex)

<CR>

Terminating carriage return

SNDF Command
SNDF is a command sent from an external host device to the ventilator system
instructing it to transmit all ventilator settings data, monitored patient data,
and alarm settings and occurrences. Enter the SNDF command exactly as
shown:
SNDF<CR>
When the ventilator receives the command SNDF<CR>, it responds with the
code MISCF, followed by ventilator settings, monitored patient data, and
alarm information
The MISCF response follows this format:
MISCF

1225*

169

<STX>

FIELD 5, FIELD 169,

<ETX>

<CR>

Response code to SNDF command

Data field, left-justified and padded


with spaces

Number of bytes between <STX> and


<CR>

End of transmission (03 hex)

Number of data fields between <STX>


and <ETX>

Terminating carriage return

Start of transmission (02 hex)

1229 if Philips is selected for serial port


in communication setup

The table below lists MISCF responses to SNDF commands

5-10

Operators Manual

Data Transfer

Note:
Non-applicable fields will either contain zero or be blank.
Table5-2.MISCF Response
Component

Operators Manual

Description

MISCF

Response to SNDF command (5 characters)

1225*

Number of bytes between <STX> and <CR> (4 characters) *1229 if Phillips is


selected for the Comm port in Communication Setup

169

Number of fields between <STX> and <ETX> (3 characters)

<STX>

Start of transmission character (02 hex)

Field 5

Ventilator time (HH:MM_) (6 characters)

Field 6

Ventilator ID to allow external hosts to uniquely identify each Puritan Bennett


980 Ventilator (18 characters)

Field 7

Date (MMM_DD_YYYY_) (12 characters)

Field 8

Vent Type (NIV______ or INVASIVE_) (9 characters)

Field 9

Mode (A/C___, SIMV__, SPONT_ or CPAP) (6 characters)

Field 10

Mandatory Type (PC____, VC____, VC+___) (6 characters)

Field 11

Spontaneous Type (PS____, TC____, VS____, PA____ (6 characters)

Field 12

Trigger Type setting (V-TRIG, P-TRIG) (6 characters)

Field 13

Respiratory rate setting in breaths/min (6 characters)

Field 14

Tidal volume (VT) setting in L (6 characters)

Field 15

Peak flow (VMAX) setting in L/min (6 characters)

Field 16

O2% setting (6 characters)

Field 17

Pressure sensitivity setting in cmH2O (6 characters)

Field 18

PEEP/CPAP in cmH2O (6 characters)

Field 19

Plateau setting in seconds (6 characters)

Field 20

Apnea interval setting in seconds (6 characters)

Field 21

Apnea tidal volume setting in L (6 characters)

Field 22

Apnea respiratory rate setting in breaths/min (6 characters)

Field 23

Apnea peak flow setting in L/min (6 characters)

5-11

Product Data Output

Table5-2.MISCF Response (Continued)


Component

5-12

Description

Field 24

Apnea O2% setting (6 characters)

Field 25

PCV apnea inspiratory pressure setting in cmH2O (6 characters)

Field 26

PCV Apnea Inspiratory Time setting in seconds (6 characters)

Field 27

Apnea flow pattern setting (SQUARE or RAMP) (6 characters)

Field 28

Apnea mandatory type setting (PC or VC) (6 characters)

Field 29

Inspiratory component of Apnea I:E ratio (if apnea mandatory type is PC)
(6 characters)

Field 30

Expiratory component of Apnea I:E ratio (if apnea mandatory type is PC)
(6 characters)

Field 31

Support pressure setting (cmH2O)

Field 32

Flow pattern setting (SQUARE or RAMP) (6 characters)

Field 33

Elevate O2 state (ON or OFF) (6 characters)

Field 34

High inspiratory pressure alarm setting (2PPEAK) in cmH2O (6 characters)

Field 35

Low inspiratory pressure alarm setting (4PPEAK) in cmH2O or OFF (6 characters)

Field 36

High exhaled minute volume (2VE TOT) alarm setting in L/min or OFF
(6 characters)

Field 37

Low exhaled minute volume (4VE TOT) alarm setting in L/min or OFF (6 characters)

Field 38

High exhaled mandatory tidal volume (2VTE MAND) alarm setting in mL or OFF
(6 characters)

Field 39

Low exhaled mandatory tidal volume (4VTE MAND) alarm setting in mL or OFF
(6 characters)

Field 40

High exhaled spontaneous tidal volume (2VTE SPONT alarm setting in mL or OFF
(6 characters)

Field 41

Low exhaled spontaneous tidal volume (4VTE SPONT) alarm setting in mL or OFF
(6 characters)

Field 42

High respiratory rate (2fTOT) alarm setting in breaths/min or OFF (6 characters)

Field 43

High inspired tidal volume (2VTI) alarm setting in mL (6 characters)

Field 44

Base flow setting in L/min (6 characters)

Field 45

Flow sensitivity (VSENS) setting in L/min (6 characters)

Operators Manual

Data Transfer

Table5-2.MISCF Response (Continued)


Component

Operators Manual

Description

Field 46

PCV inspiratory pressure (PI) setting in cmH2O (6 characters)

Field 47

PCV inspiratory time (TI) setting in seconds (6 characters)

Field 48

Inspiratory component of I:E ratio setting or High component of H:L ratio setting
(6 characters)

Field 49

Expiratory component of I:E ratio setting or Low component of H:L ratio setting
(6 characters)

Field 50

Constant during rate change setting (I-time, I/E, or E-time) (6 characters)

Field 51

Tube ID setting in mm (6 characters)

Field 52

Tube Type setting (ET or TRACH) (6 characters)

Field 53

Humidification type setting (Non-heated exp tube, Heated exp tube, or HME)
(18 characters)

Field 54

Humidifier volume setting in L (6 characters)

Field 55

O2sensor setting (Enabled or Disabled) (9 characters)

Field 56

Disconnect sensitivity (DSENS) setting in % or OFF (6 characters)

Field 57

Rise time % setting (6 characters)

Field 58

PAV+ percent support setting (6 characters)

Field 59

Expiratory sensitivity (ESENS) setting in % or L/min for PAV+ breath type


(6 characters)

Field 60

PBW setting in kg (6 characters)

Field 61

Target support volume (VT SUPP) setting in L (6 characters)

Field 62

High pressure (PH) setting (in BiLevel) in cmH2O (6 characters)

Field 63

Low pressure (PL) setting (in BiLevel) in cmH2O (6 characters)

Field 64

High pressure time (TH) setting (in BiLevel) in seconds (6 characters)

Field 65

High spontaneous inspiratory time limit (2TI SPONT) setting in seconds


(6 characters)

Field 66

Circuit type setting (ADULT, PEDIATRIC, or NEONATAL) (9 characters)

Field 67

Low pressure time (TL) setting (in BiLevel) in seconds (6 characters)

Field 68

Expiratory time (TE) setting in seconds (6 characters)

Field 69

End inspiratory pressure (PI END) in cmH2O (6 characters)

5-13

Product Data Output

Table5-2.MISCF Response (Continued)


Component

5-14

Description

Field 70

Respiratory rate (fTOT) in breaths/min (6 characters)

Field 71

Exhaled tidal volume (VTE ) in L (6 characters)

Field 72

Patient exhaled minute volume (VE TOT) in L/min (6 characters)

Field 73

Peak airway pressure (PPEAK) in cmH2O (6 characters)

Field 74

Mean airway pressure (PMEAN) in cmH2O (6 characters)

Field 75

Expiratory component of monitored value of I:E ratio, assuming inspiratory component of 1 (6 characters)

Field 76

I:E ratio (6 characters)

Field 77

Delivered O2% (6 characters)

Field 78

Inspired tidal volume (VTI) in L (6 characters)

Field 79

Intrinsic PEEP (PEEPI) in cmH2O (6 characters)

Field 80

Estimated total resistance (RTOT) in cmH2O/L/s (6 characters)

Field 81

Estimated patient resistance (RPAV) in cmH2O/L/s (6 characters)

Field 82

Estimated patient elastance (EPAV) in cmH2O/L (6 characters)

Field 83

Estimated patient compliance (CPAV) in mL/cmH2O (6 characters)

Field 84

Not used

Field 85

Rapid shallow breathing index (f/VT) (6 characters)

Field 86

Spontaneous percent inspiratory time (TI/TTOT) (6 characters)

Field 87

Monitored PEEP in cmH2O (6 characters)

Field 88

Spontaneous inspiratory time (TI SPONT) in seconds (6 characters)

Field 89

Exhaled spontaneous minute volume (VE SPONT) in L/min (6 characters)

Field 90

Intrinsic PEEP (PEEPI) from expiratory pause maneuver in cmH2O (6 characters)

Field 91

Total PEEP (PEEPTOT) from expiratory pause maneuver in cmH2O (6 characters)

Field 92

Static compliance (CSTAT) from inspiratory pause maneuver in mL/cmH2O


(6 characters)

Field 93

Static resistance (RSTAT) from inspiratory pause maneuver in cmH2O/L/s


(6 characters)

Operators Manual

Data Transfer

Table5-2.MISCF Response (Continued)


Component

Description

Field 94

Plateau pressure (PPL) from inspiratory pause maneuver in cmH2O (6 characters)

Field 95

High spontaneous inspiratory time (ALERT_ or blank)


(6 characters)

Field 96

Dynamic compliance (CDYN) in mL/cmH2O (6 characters)

Field 97

Dynamic resistance (RDYN) in cmH2O/L/s (6 characters)

Field 98

Peak spontaneous flow (PSF) in L/min (6 characters)

Field 99

Peak expiratory flow (PEF) in L/min (6 characters)

Field 100

End expiratory flow (EEF) in L/min (6 characters)

Field 101

Proximal Flow Sensor state) ON or OFF)

Field 102

Negative inspiratory force (NIF) in cmH2O (6 characters)

Field 103

P0.1 pressure change in cmH2O (6 characters)

Field 104

Vital capacity (VC) in L (6 characters)

Field 105

Alarm Silence (ON or OFF) (6 characters)

Field 106

Apnea ventilation alarm* (6 characters)

Field 107

High exhaled minute volume alarm* (1VE TOT) (6 characters)

Field 108

High exhaled tidal volume alarm* (1VTE ) (6 characters)

Field 109

High O2% alarm* (6 characters)

Field 110

High inspiratory pressure alarm* (1PPEAK) (6 characters)

Field 111

High ventilator pressure alarm* (1PVENT) (6 characters)

Field 112

High respiratory rate alarm* (1fTOT) (6 characters)

Field 113

AC power loss alarm* (6 characters)

Field 114

Inoperative battery alarm* (6 characters)

Field 115

Low battery alarm* (6 characters)

Field 116

Loss of power alarm* (6 characters)

Field 117

Low exhaled mandatory tidal volume alarm* (3VTE MAND) (6 characters)

Field 118

Low exhaled minute volume alarm* (3VE TOT) (6 characters)

* Possible responses are: NORMAL, LOW, MEDIUM, HIGH, or RESET.

Operators Manual

5-15

Product Data Output

Table5-2.MISCF Response (Continued)


Component

Description

Field 119

Low exhaled spontaneous tidal volume (3VTE SPONT) alarm* (6 characters)

Field 120

Low O2% alarm* (6 characters)

Field 121

Low air supply pressure alarm* (6 characters)

Field 122

Low O2 supply pressure alarm* (6 characters)

Field 123

Reserved

Field 124

Disconnect alarm* (6 characters)

Field 125

Severe occlusion alarm* (6 characters)

Field 126

Inspiration too long alarm* (6 characters)

Field 127

Procedure error* (6 characters)

Field 128

Compliance limited tidal volume (VT) alarm* (6 characters)

Field 129

High inspired spontaneous tidal volume* (3TI SPONT) alarm (6 characters)

Field 130

High inspired volume* (1VTI) alarm (6 characters)

Field 131

High compensation limit (1PCOMP) alarm* (6 characters)

Field 132

PAV+ startup too long alarm* (6 characters)

Field 133

PAV+ R and C not assessed alarm* (6 characters)

Field 134

Volume not delivered (VC+) alarm* (6 characters)

Field 135

Volume not delivered (VS) alarm* (6 characters)

Field 136

Low inspiratory pressure (3PPEAK) alarm* (6 characters)

Field 137

Technical malfunction A5* (6 characters)

Field 138

Technical malfunction A10* (6 characters)

Field 139

Technical malfunction A15* (6 characters)

Field 140

Technical malfunction A20* (6 characters)

Field 141

Technical malfunction A25* (6 characters)

Field 142

Technical malfunction A30* (6 characters)

Field 143

Technical malfunction A35* (6 characters)

Field 144

Technical malfunction A40* (6 characters)

* Possible responses are: NORMAL, LOW, MEDIUM, HIGH, or RESET.

5-16

Operators Manual

Data Transfer

Table5-2.MISCF Response (Continued)


Component

Description

Field 145

Technical malfunction A45* (6 characters)

Field 146

Technical malfunction A50* (6 characters)

Field 147

Technical malfunction A55* (6 characters)

Field 148

Technical malfunction A60* (6 characters)

Field 149

Technical malfunction A65* (6 characters)

Field 150

Technical malfunction A70* (6 characters)

Field 151

Technical malfunction A75* (6 characters)

Field 152

Technical malfunction A80* (6 characters)

Field 153

Technical malfunction A85* (6 characters)

Field 154

Spontaneous tidal volume (VTE SPONT) in liters (6 characters)

Field 155

Total work of breathing (WOBTOT) in Joules/L (6 characters)

Field 156

Leak Sync state (9 characters) (ON, or OFF)

Field 157

%LEAK (6 characters)

Field 158

LEAK (6 characters)

Field 159

VLEAK (6 characters)

Field 160

Prox Inop alarm* (ALARM or NORMAL)

Field 161 - 171

Reserved

<ETX>

End of transmission character (03 hex)

<CR>

Terminating carriage return

* Possible responses are: NORMAL, LOW, MEDIUM, HIGH, or RESET.

5.5

Communication Ports
WARNING:
To avoid possible injury, only connect devices that comply with IEC 60601-1
standard to any of the ports at the rear of the ventilator, with the exception
of passive memory storage devices (flash drives) and serial-to-USB adapter
cables. If a serial-to-USB adapter cable is used, it must be connected to an IEC
60601-1-compliant device.

Operators Manual

5-17

Product Data Output

WARNING:
To avoid possible injury, do not connect a device that is attached to the
patient to any of the non-clinical ports listed below when the ventilator is
ventilating a patient.

Figure5-2.Port Locations

5-18

RS-232 Port (COM 1)

Ethernet Port (non-clinical use)

Nurse call Port (remote alarm port)

USB Port (USB 1) (COM 2) (non-clinical use)

USB Port (USB 2) (COM 3) (non-clinical use)

HDMI Port (non-clinical use)

Service Port (non-clinical use)

Operators Manual

Data Transfer

5.5.1

Port Use
Reference Data Transfer, p. 5-1 for data transfer details.
RS-232 Port
To use the RS-232 port
1.
Obtain a cable with a male DB-9 connector to connect to the RS-232 port on the
ventilator.
2.

Make the appropriate connection to a monitoring device. A gender changer, null


modem cable or socket saver may be required. Consult with the institutions Information Technology professional as required.

3.

Ensure to specify the baud rate, parity, and data bits in the ventilator communication setup to correctly match the parameters of the monitoring device.

4.

A monitor designed to use this port is required for obtaining data from the ventilator. Set up the monitoring device to receive ventilator data. These data can
include waveform data.

5.

Program the remote device to send the appropriate RS-232 commands as


described in the next section.

Reference Reference MISCA Response, p. 5-6 and Reference MISCF Response,


p. 5-11 for MISCA and MISCF responses to SNDA and SNDF commands.
Ethernet Port
The Ethernet port is used by Service personnel for accessing various logs and
updating ventilator software.
Nurse Call Port
A remote alarm or nurse call interface is available on the ventilator system
which can be used to remotely annunciate the alarm status of the ventilator.
Medium and high priority alarms are remotely annunciated. The nurse call connector is located at the back of the ventilator, as shown. Reference Port Locations, p. 5-18.
Reference the remote alarm manufacturers instructions for use for information regarding proper nurse call connection.

Operators Manual

5-19

Product Data Output

USB Ports
The USB ports are used for screen captures, or receiving serial data when a USB
port has been configured as a serial port. This is also known as transferring
data via a serial-over-USB protocol. Reference Communication Setup, p. 5-3
for Comm setup configuration. Screen captures require an external USB
memory storage device (flash drive) for screen captures. Instructions for
using this port for screen captures are given. Reference To capture GUI screens,
p. 5-2.
HDMI Port
An external display can be used via connection with the HDMI port.
To use the HDMI port with an external display
1.
Connect one end of an HDMI cable to the HDMI port at the back of the ventilator
(item 6, above).
2.

Connect the other end of the cable to the external display. An HDMI to DVI
adapter may be used.

3.

Turn the device on. The appearance of the GUI now displays on the external
display device.

Service Port
The Service port is used by service personnel only.
5.6

Retrieving Stored Data


Ventilator data are stored in various logs, accessible using the logs icon. Some
logs may be accessed during normal ventilation, and some are only available
to Covidien personnel when the ventilator is in Service mode. Reference Ventilator Logs, p. 8-2 for more information on data stored in various logs.

5.7

Display Configurability
The operator can configure some ventilator parameters according to personal
preference. Reference Ventilator Configuration, p. 3-35 for a table showing
which parameters are configurable and by whom.

5-20

Operators Manual

Data Transfer

Reference Preparing the Ventilator for Use, p. 3-37 for information on configuring each display item.
5.8

Printing Data or Screen Captures


The ventilator cannot be connected directly to a printer.
Save screen captures to an external storage device, such as a USB flash drive,
then print from a PC. Reference GUI Screen Capture, p. 5-2 for instructions on
using the screen-capture feature.

5.9

Connectivity to External Systems


The ventilator is compatible with the Philips Medical IntelliVue MP50 and
Spacelabs Ultraview patient monitoring systems.

Note:
Not all patient monitors are compatible with the Puritan Bennett 980 Series
Ventilator.

Operators Manual

5-21

Product Data Output

Page Left Intentionally Blank

5-22

Operators Manual

6 Performance

6.1

Overview
This chapter contains detailed information about Puritan Bennett 980 Series
Ventilator performance including:

6.2

Ventilator settings

Alarm interpretation and alarm testing

A detailed description of selected alarms

Monitored patient data

System Options
Various software options are available for the ventilator. Details for each of
these options are described in the appendices included in this manual.

6.3

6.4

Environmental Considerations
WARNING:
Use of the ventilator in altitudes higher or barometric pressures lower than
those specified could compromise ventilator operation. Reference
Environmental Specifications, p. 11-7 for a complete list of environmental
specifications.

Ventilator Settings
Default ventilator settings are based on the circuit type selected during SST. A
neonatal, pediatric or adult patient circuit can be used, and all accessories
needed to ventilate the patient should be attached when SST is performed.

6-1

Performance

6.4.1

Ventilation Type
The clinician enters the vent type, specifying how the patient will be ventilated;
invasively or non-invasively (NIV). The vent type optimizes the alarm limits for
NIV patients, and disables some settings for NIV ventilation.

6.4.2

Mode
Available ventilation modes are mandatory (A/C) or spontaneous (SPONT)
modes, as well as two mixed modes: SIMV and BiLevel.

A/C (Assist-Control) A/C mode guarantees delivery of a minimum number of


mandatory breaths based on the frequency (f) set by the clinician. Breaths in A/C
can be patient-initiated (PIM) or ventilator-initiated (VIM).

SPONT (Spontaneous) SPONT mode delivers only spontaneous breaths


which are all patient-initiated.

SIMV (Synchronized Intermittent Mandatory Ventilation) SIMV is a


mixed mode allowing both mandatory and spontaneous breaths. SIMV guarantees at least one mandatory breath per set breath cycle, which is either patientinitiated or ventilator-initiated. The mandatory type of an SIMV breath can be PC,
VC, or VC+.

BiLevel BiLevel is also a mixed mode which overlays the patients spontaneous breaths onto the breath structure for PC mandatory breaths. Two levels of
pressure, PL and PH are employed. The breath cycle interval for both SIMV and
BiLevel modes is 60/f where f is the respiratory rate set by the operator.

CPAP CPAP is available only when circuit type is neonatal and vent type is NIV.

CPAP mode allows spontaneous breathing with a desired PEEP level. In


order to limit inadvertent alarms associated with the absence of returned
volumes in nasal CPAP breathing, CPAP does not make available exhaled
minute volume and exhaled tidal volume alarm settings.
6.4.3

Breath Type
Mandatory breath types for A/C and SIMV modes include volume controlled
(VC), pressure controlled (PC), or volume control plus (VC+) breath types, also
called Mandatory Type.

6-2

VC (Volume Control) The ventilator delivers an operator-set tidal volume.

PC (Pressure Control) The ventilator delivers an operator-set pressure.

VC+ (Volume Control Plus) Volume control plus (a mandatory, pressure


controlled breath type that does not restrict flow during the inspiratory phase, and

Operators Manual

Ventilator Settings

automatically adjusts the inspiratory pressure target from breath to breath to


achieve the desired tidal volume despite changing lung conditions. Reference
Mandatory Breath Delivery, p. 10-17 for more information on VC+.

Mandatory inspirations are triggered in the following ways

Pressure Trigger (P-TRIG) Changes in circuit pressure cause the ventilator to


deliver a breath. These pressure changes relate to the pressure sensitivity (PSENS)
set by the operator. If the patient makes an effort to inspire, the airway pressure
drops. If the pressure drops by at least the value of PSENS, the ventilator delivers a
breath.

Flow Trigger (V-TRIG) Changes in flow in the circuit cause the ventilator to
deliver a breath. The breath delivery and exhalation valve flow sensors measure
gas flow in the ventilator breathing system. As the patient inspires, the delivered
flow remains constant and the exhalation valve flow sensor measures decreased
flow. When the difference between the two flow measurements is at least the
operator-set value for flow sensitivity (VSENS), the ventilator delivers a breath.

Time Trigger The ventilator delivers a Ventilator Initiated Mandatory (VIM)


breath after a specific amount of time elapses.

Operator Trigger (OIM) The operator presses the Manual inspiration key.
An operator initiated mandatory breath is also called an OIM breath. During an
OIM breath, the breath delivered is based on the current settings for a mandatory
breath.

Spontaneous breathing modes such as SIMV, BiLevel, and SPONT include the
following breath types (called Spontaneous Types):

PS (Pressure Support) The ventilator delivers an operator-set positive pressure above PEEP (or PL in BiLevel) during a spontaneous breath. If SIMV is selected
as the mode, PS is automatically selected for spontaneous type.

VS (Volume Support) The ventilator delivers an operator-set positive pressure above PEEP during a spontaneous breath and automatically adjusts the pressure level from breath to breath to consistently deliver the set tidal volume.

TC (Tube Compensation) Additional positive pressure delivered to the


patient during spontaneous breaths to overcome resistance of the artificial airway.

PAV+ (Proportional Assist Ventilation)) A software option that allows the


ventilator to reduce the work of breathing (WOB) by assisting the patients inspiration by an operator-set amount proportional to the breathing effort generated
by the patient. Reference Appendix D for more information on PAV+.

The inspiratory trigger methods for spontaneous breaths are

Operators Manual

Pressure Trigger (P-TRIG) Same as described for mandatory inspiration triggers.

6-3

Performance

Flow Trigger (V-TRIG) Same as described for mandatory inspiration triggers.

IE Sync Trigger When this software option is installed and the IE Sync trigger
type is selected, triggering and cycling of spontaneous breaths occur based on the
patients intra-pleural pressure.

Operator Trigger (OIM) Since the operator can only initiate a mandatory
breath by pressing the Manual inspiration key, spontaneous mode allows OIMs,
but the breath delivered is based on the current apnea breath settings.

Reference Inspiration Detection and initiation, p. 10-4 for details on the different trigger methods.
6.5

Alarms
This manual uses the following conventions when discussing alarms:
A description or name of an alarm without specifying the alarm setting is
denoted with an upward or downward pointing arrow (1 or 3) preceding the
specific alarm name. An alarm setting is denoted as an upward or downward
pointing arrow with an additional horizontal limit symbol (2 or 4) preceding the
specific alarm. Some alarm conditions actually limit breath delivery such as
1PPEAK and 1VTI by truncating inspiration and transitioning to the exhalation
phase. These alarm conditions are denoted as alarm limits. Reference Alarm
Descriptions and Symbols, p. 6-6.

6.5.1

Alarm Messages
Alarms are visually annunciated using an indicator on the top of the GUI, which
has a 360 field of view. If an alarm occurs, this indicator flashes at a frequency
and color matching the alarm priority. The alarms also appear as colored
banners on the right side of the GUI screen. If an alarm occurs, this indicator
appears in the color matching the alarm priority (yellow for low (!) and medium
(!!) priority; red for high (!!!) priority. For technical alarm and non-technical
alarm details, reference the respective tables on p. 6-18 and p. 6-29.
An alarm is defined as a primary alarm if it is the initial alarm. A dependent
alarm arises as a result of conditions that led to the primary alarm. This is also
referred to as an augmentation. An augmentation strategy is built into the ventilator software to handle occurrences where the initial cause of the alarm has
the potential to precipitate one or more additional alarms. When an alarm
occurs, any subsequent alarm related to the cause of this initial alarm aug-

6-4

Operators Manual

Ventilator Settings

ments the initial alarm instead of appearing on the GUI as a new alarm. The
initial alarms displayed analysis message is updated with the related alarms
information, and the Alarm Log Event column shows the initial alarm as Augmented.
A primary alarm consists of a base message, analysis message, and a
remedy message. The base message describes the primary alarm. The analysis message describes the likely cause of the alarm and may include alarm augmentations. The remedy message provides information on what to do to
correct the alarm condition.
Alarm banners, when dragged leftward from the right side of the GUI, display
messages for the indicated active alarms. The figure below, shows the alarm
message format.
Figure6-1.Alarm Message Format

Base message

Analysis message

Remedy message

A latched alarm is one whose visual alarm indicator remains illuminated even
if the alarm condition has autoreset. Latched alarm indicators are located on
the sides of the omni-directional LED. A latched alarm can be manually reset
by pressing the alarm reset key. If no alarms are active, the highest priority

Operators Manual

6-5

Performance

latched alarm appears on the omni-directional LED on the GUI. A lockable


alarm is one that does not terminate an active alarm silence function (it does
not sound an audible alert during an active alarm silence function), while a
non-lockable alarm cancels the alarm silence period and sounds an audible
alert. All patient data alarms and the CIRCUIT DISCONNECT alarm are lockable
alarms.

Note:
When a new lockable alarm occurs, the alarm will not start to sound audibly if the
previous lockable alarm was silenced.

The following rules define how alarm messages are displayed:

Primary alarms precede any dependent alarms.

The system adds dependent alarms to the analysis messages of each active
primary alarm with which they are associated. If a dependent alarm resets, the
system removes it from the analysis message of the primary alarm.

The priority level of a primary alarm is equal to or greater than the priority level of
any of its active dependent alarms.

An alarm cannot be a dependent alarm of any alarm that occurs subsequently.

If a primary alarm resets, any active dependent alarms become primary unless they
are also dependent alarms of another active primary alarm. This is due to different
reset criteria for primary and dependent alarms.

The system applies the new alarm limit to alarm calculations from the moment a
change to an alarm limit is accepted.

The priority level of a dependent alarm is based solely on its detection conditions
(not the priority of any associated alarms.

When an alarm causes the ventilator to enter OSC or safety valve open (SVO), the
patient data display (including waveforms) is blanked. The elapsed time without
ventilatory support (that is, since OSC or SVO began) appears on the GUI screen.
If the alarm causing OSC or SVO is autoreset, the ventilator resets all patient data
alarm detection algorithms.

Table6-1.Alarm Descriptions and Symbols


Alarm description

6-6

Symbol

High compensation pressure

1PCOMP

High delivered oxygen percentage

1O2%

Operators Manual

Ventilator Settings

Table6-1.Alarm Descriptions and Symbols (Continued)


Alarm description
High exhaled minute volume

1VE TOT

High exhaled minute volume setting

2VE TOT

High exhaled tidal volume

1VTE

High exhaled tidal volume setting

2VTE

High inspired tidal volume limit

2VTI

High internal ventilator pressure

1PVENT

High respiratory rate

1fTOT

High respiratory rate setting

2fTOT

High spontaneous inspiratory time

1TI SPONT

High spontaneous inspiratory time limit

2TI SPONT

High circuit pressure

1PPEAK

High circuit pressure limit

2PPEAK

Low circuit pressure

3PPEAK

Low circuit pressure setting

4PPEAK

Low exhaled mandatory tidal volume

3VTE MAND

Low exhaled mandatory tidal volume


setting

4VTE MAND

Low exhaled minute volume

3VE TOT

Low exhaled minute volume setting

4VE TOT

Low exhaled spontaneous tidal volume

3VTE SPONT

Low exhaled spontaneous tidal volume


setting

4VTE SPONT

Low delivered oxygen percentage

Operators Manual

Symbol

3O2%

6-7

Performance

6.5.2

Alarm Reset Key


The alarm reset function can be used for any non-technical alarm. Reference
Alarm Handling, p. 6-15 for an explanation of technical vs. non-technical
alarms. Alarm reset reinitializes the algorithm the ventilator uses to initially
detect the alarm except for A/C POWER LOSS, LOW BATTERY, NO AIR SUPPLY,
NO O2 SUPPLY, PROCEDURE ERROR alarms and active battery alarms. If the
cause of the alarm still exists after the Alarm Reset key is pressed, the alarm
becomes active again. The ventilator logs all actuations of the alarm reset key.

6.5.3

Alarm Silence Key


The alarm silence feature temporarily mutes the audible portion of an alarm
for two minutes. After the two-minute period, if the alarm condition still exists,
the alarm sounds again. Pressing the alarm silence key again re-starts the two
minute interval during which an alarm is muted. An LED within the key illuminates and a count-down timer appears on the GUI next to an alarm silence
indicator symbol, indicating an active alarm silence function. The alarm silence
feature does not allow the audible alarm to be turned off; the audible portion
of the alarm is temporarily muted for two minutes. The GUIs omni-directional
LED flashes during an active alarm state and during an alarm silence period and
its appearance changes with the priority if the alarm escalates. Pressing the
Alarm Reset key cancels an Alarm Silence. If the condition that caused the
alarm still exists, the alarm activates again.

6.5.4

Alarm Volume Key


An alarm volume key is available for setting the desired alarm volume. The
alarm volume is automatically set to the factory default setting of 10 (maximum) or to the institutional default setting based on circuit type if it has been
so configured. When setting the alarm volume, a sample tone is generated,
allowing the practitioner to decide the appropriate alarm volume for the surrounding ambient conditions. If a high priority alarm occurs, the alarm volume
increases one (1) increment from its current volume level if it is not acknowledged within 30 s. If a high priority alarm is not acknowledged within 60 s, the
audible alarm volume escalates to its maximum volume.
Reference To adjust alarm volume, p. 3-41 for instructions on adjusting the
alarm volume.

6-8

Operators Manual

Ventilator Settings

6.5.5

WARNING:
The audio alarm volume level is adjustable. The operator should set the
volume at a level that allows the operator to distinguish the audio alarm
above background noise levels. Reference To adjust alarm volume, p. 3-41.

Alarm Testing
Testing the alarms requires oxygen and air sources and stable AC power. Test
the alarms at least every six months, using the procedures described.
Required Equipment

Test lung (P/N 4-000612-00)

Adult patient circuit

If the alarm does not annunciate as indicated, verify the ventilator settings and
repeat the test. The alarm tests check the operation of the following alarms:

CIRCUIT DISCONNECT

LOW EXHALED MANDATORY TIDAL VOLUME (3VTE MAND)

LOW EXHALED TOTAL MINUTE VOLUME (3VE TOT)

HIGH CIRCUIT PRESSURE (1PPEAK)

SEVERE OCCLUSION

AC POWER LOSS

APNEA

LOW EXHALED SPONTANEOUS TIDAL VOLUME (3VTE SPONT)

NO O2 SUPPLY

LOW DELIVERED O2% (3O2%)

HIGH DELIVERED O2% (1O2%)

Ventilator setup for alarms tests


1.

Disconnect the patient circuit from the ventilator and turn the ventilator off for at
least five minutes.

2.

Turn the ventilator or on. The ventilator runs POST.

Operators Manual

6-9

Performance

3.

On the GUI, select NEW PATIENT.

4.

Set up new patient using the following settings.


PBW: 70 kg
Vent type: INVASIVE
Mode: A/C
Mandatory type: VC
Trigger type:V-TRIG

5.

Set the following new patient settings


f: 6.0 1/min
VT: 500 mL
VMAX:

30 L/min

TPL: 0 s
Flow pattern: SQUARE
VSENS:

3 L/min

O2%: 21%
PEEP: 5 cmH2O
6.

Set the following apnea settings


TA: 10 s
f: 6.0 1/min
O2%: 21%
VT: 500 mL

7.

Set the following alarm settings


2PPEAK: 70 cmH2O
fTOT: OFF
4VE TOT: 1 L/min
2VE TOT: 3.5 L/min
4VTE MAND: 300 mL
2VTE MAND: OFF
4VTE SPONT: OFF
2VTE SPONT: OFF

6-10

Operators Manual

Ventilator Settings

8.

Set the graphics display to a volume-time plot (for use in the APNEA alarm test).

9.

Connect an Adult patient circuit to the ventilator and attach a test lung to the
patient wye.

Note:
To ensure proper test results, do not touch the test lung or patient circuit during the
CIRCUIT DISCONNECT alarm test.
CIRCUIT DISCONNECT alarm test
1.
Allow the ventilator to deliver at least four breaths. During the inspiratory phase
of a breath, disconnect the inspiratory filter from the To Patient port. The ventilator annunciates a CIRCUIT DISCONNECT alarm after the inspiratory filter is disconnected.
2.

Connect the inspiratory filter to the To Patient port to autoreset the alarm.

LOW EXHALED MANDATORY TIDAL VOLUME (3VTE MAND) alarm test


Set VT to 225 mL. The ventilator annunciates a LOW EXHALED MANDATORY
TIDAL VOLUME (3VTE MAND) alarm on the third consecutive breath after Accept
is touched.
LOW EXHALED TOTAL MINUTE VOLUME (3VE TOT) alarm test
Set 4VE TOT alarm limit to 3.45 L/min. The ventilator annunciates a LOW
EXHALED TOTAL MINUTE VOLUME (3VE TOT) alarm on the next breath after
Accept is touched.
HIGH CIRCUIT PRESSURE (1PPEAK) alarm test
1.

Make the following patient and alarm settings changes.


VT: 500 mL
VMAX:

30 L/min

2PPEAK: 20 cmH2O
2.

After one breath, the ventilator annunciates a HIGH CIRCUIT PRESSURE (1PPEAK)
alarm. If the alarm does not sound, check the patient circuit for leaks.

SEVERE OCCLUSION alarm test


1.
Make the following alarm settings changes.
2PPEAK: 50 cmH2O

Operators Manual

6-11

Performance

2.

Press the alarm reset key to reset all alarms.

3.

Adjust DSENS to the VMAX setting.

4.

Disconnect the ventilator breathing circuit from the FROM PATIENT port and block
the gas flow.

5.

While maintaining the occlusion, ensure the safety valve open indicator appears
on the status display, the GUI shows the elapsed time without normal ventilation
support, and the test lung inflates and deflates rapidly with small pulses as the
ventilator delivers trial pressure-based breaths.

6.

Press the alarm reset key to reset all the alarms.

AC POWER LOSS alarm test


1.
Allow the ventilator to deliver at least four breaths, then disconnect the power
cord from AC facility power. If any battery is charged, the GUI annunciates an AC
POWER LOSS alarm. If less than ten minutes of battery backup are available, the
GUI annunciates a LOW BATTERY alarm. If no battery power is available, the BDU
annunciates a LOSS OF POWER alarm.
2.

Connect the power cord to AC facility power. The AC POWER LOSS or LOW
BATTERY alarm autoresets.

APNEA alarm test


1.
Make the following alarm settings changes:
2PPEAK: 70 cmH2O
Mode: SPONT
Spontaneous type: PS

Note:
To avoid triggering a breath during the apnea interval, do not touch the test lung
or patient circuit.

Note:
For the apnea alarm test, the exhaled tidal volume (VTE) displayed in the patient
data area must be greater than half the delivered volume shown on the volumetime plot in the graphics display in order for apnea to autoreset. Reference Apnea
Ventilation, p. 10-41 for a technical description of apnea ventilation.
2.

6-12

The GUI annunciates an APNEA alarm within 10 s after touching Accept.

Operators Manual

Ventilator Settings

3.

Squeeze the test lung twice to simulate two subsequent patient-initiated breaths.
The APNEA alarm autoresets.

4.

Let the ventilator return to apnea ventilation.

LOW EXHALED SPONTANEOUS TIDAL VOLUME alarm test


1.
Make the following patient and alarm settings changes
Trigger type: P-TRIG
2PSENS: 4 cmH2O
4VTE SPONT: 2500 mL
2.

Press the alarm reset key to reset the apnea alarm.

3.

Slowly squeeze the test lung to simulate spontaneous breaths. The ventilator
annunciates a LOW EXHALED SPONTANEOUS TIDAL VOLUME (3VTE SPONT) alarm
at the start of the fourth consecutive spontaneous inspiration.

4.

Make the following patient settings changes:


Mode: A/C
4VTE SPONT: OFF

5.

Press the alarm reset key to reset the 4VTE SPONT alarm.

NO O2 SUPPLY alarm test


1.

Disconnect the oxygen inlet supply. The ventilator annunciates a NO O2 SUPPLY


alarm within one breath.

2.

Connect the oxygen inlet supply. The NO O2 SUPPLY alarm autoresets within two
breaths after oxygen is reconnected.

LOW DELIVERED O2% and HIGH DELIVERED O2% alarms tests


1.

Make the following patient and alarm settings changes:


PSENS: 2 cmH2O
O2%: 100%

2.

Make the following apnea settings changes:


TA: 60 s

3.

Operators Manual

Attach the ventilators oxygen gas hose to a known air supply (for example, a
medical grade air cylinder) or a wall air outlet.

6-13

Performance

6.5.6

4.

attach the ventilators air gas hose to a known medical oxygen supply.

5.

Observe the GUI screen. The delivered O2% display should decrease, and the ventilator should annunciate a medium priority 3O2% alarm within 60 s and a high
priority 3O2% alarm within two (2) minutes.

6.

Set the O2% to 21%.

7.

Observe the GUI screen. The delivered O2% display should increase, and the ventilator should annunciate a a medium priority 1O2% alarm within 60 s and a high
priority 1O2% alarm within two (2) minutes.

8.

Remove the air gas hose from the oxygen supply and reconnect the hose to a
known medical air supply.

9.

Remove the oxygen gas hose from the air supply and reconnect the hose to a
known oxygen supply.

10.

Press the alarm reset key to clear all alarms.

WARNING:
Before returning the ventilator to service, review all settings and set
appropriately for the patient to be ventilated.

Viewing Alarms
When an alarm occurs, the omni-directional LED at the top of the GUI flashes
in a color corresponding to the alarm priority, an audible series of tones
sounds, and an alarm banner displays on the GUI. Reference Areas of the GUI,
p. 4-3. When the alarm banner appears, it displays its base message. Touching
the individual alarm causes an expanded explanation to appear, containing
analysis and remedy messages, and may contain a link to the alarm log or the
alarms settings screen. Touch the link to display requested information. The
omni-directional LED remains steadily lit and may appear multicolored,
meaning that multiple alarms with varying priority levels have occurred. During
an event that causes multiple alarms, the ventilator simultaneously displays the
two highest priority active alarms.

6-14

Operators Manual

Ventilator Settings

6.5.7

Alarm Delay
Determination of an Alarm Condition
The delay time from the moment the alarm condition first occurs until the
alarm is annunciated is imperceptible.
Delay to/from a Distributed Alarm System
For alarm conditions relayed via the serial port, the overall delay is dependent
upon the polling rate of the external device. The delay from the time the serial
port is polled by the external device, until the alarm message leaves the serial
port does not exceed 3 seconds. An example of an external device is a patient
monitor.

6.5.8

Alarm Handling
Current alarm settings are saved in the ventilators non-volatile memory
(NVRAM). If the alarm settings are changed by another clinician, those settings become applicable. For example, there are no operator-selectable default
alarm settings.
The ventilator systems alarm handling strategy is intended to

Detect and call attention to legitimate causes for caregiver concern as quickly as
possible, while minimizing nuisance alarms.

Identify the potential cause and suggest corrective action for certain types of
alarms. However, the clinician must make the final decision regarding any clinical
action.

Make it easy to discern an alarms priority level.

Allow quick and easy alarm setup.

Ventilator alarms are categorized as high priority, medium priority, or low priority, and are classified as technical or non-technical.
The ventilator is equipped with two alarms the primary alarm and secondary
alarm. The primary alarm annunciates high, medium, and low priority alarms
when they occur. The secondary alarm (also named immediate priority in the
table below) is a continuous tone alarm and annunciates during Vent Inop conditions or complete loss of power. This alarm is powered by a capacitor and
lasts for at least 120 seconds.

Operators Manual

6-15

Performance

The table below lists alarm priority levels and their visual, audible, and autoreset characteristics. An alarm autoresets when the condition causing the alarm
no longer exists.
Table6-2.Alarm Prioritization

6-16

Priority Level

Visual indicator

Audible indicator

Autoreset
characteristics

Immediate

Specific to alarm condition or component


failure.

Continuous tone
alarm sounding for at
least 120 s.

N/A

High: Immediate attention required to ensure


patient safety.

Flashing red omnidirectional LED located


on the top of the GUI,
red alarm banner on
GUI screen, red bar
next to alarm setting
icon on Alarms screen.

High-priority audible
alarm (a sequence of
five tones that repeats
twice, pauses, then
repeats again).

Visual alarm does not


auto reset. Visual
alarm indicators
remain steadily illuminated following an
autoreset. The alarm
reset key must be
pressed to extinguish
visual indicator.

Medium: Prompt
attention necessary.

Flashing yellow omnidirectional LED located


on the top of the GUI,
yellow alarm banner
on GUI screen, and
yellow bar next to
alarm setting icon on
Alarms screen.

Medium-priority
audible alarm (a
repeating sequence of
three tones).

LED indicator turns off


and autoreset is
entered into the alarm
log.

Low: A change in the


patient-ventilator
system has occurred.

Steadily illuminated
yellow omni-directional LED located on the
top of the GUI, yellow
alarm banner on GUI
screen, and yellow bar
next to alarm setting
icon on Alarms screen.

Low-priority audible
alarm (two tone, nonrepeating).

LED indicator turns off


and autoreset is
entered into the alarm
log.

Operators Manual

Ventilator Settings

Table6-2.Alarm Prioritization (Continued)


Priority Level

Visual indicator

Audible indicator

Autoreset
characteristics

Normal: Normal ventilator operation

Steadily illuminated
green omni-directional
LED located on the top
of the GUI, no alarm
banner, and white
values next to alarm
setting icon on Alarms
screen.

None.

None

Immediate

Status display shows


the GUI has failed.

The secondary alarm


annunciates a repeating sequence of single
tones, since the
primary alarm (part of
the GUI) has failed.

None

A technical alarm is one that is caused by a violation of any of the ventilators


self monitoring conditions, such as failure of POST or a fault detected by the
ventilators background diagnostic system. Technical alarms cannot be reset by
pressing the Alarm Reset key. (Reference Background Diagnostic System, p.
10-74). Technical alarms fall into eight categories, shown in the table below.
Table6-3.Technical Alarm Categories
Category

Name

Priority

System Response

Vent-Inop

High

Ventilator goes to safe state. Reference Ventilator Protection Strategies, p. 4-35.

Exh BUV

High

Backup ventilation

Insp BUV

High

Backup ventilation

Mix BUV

High

Backup ventilation

High

Ventilator goes to safe state. Reference Ventilator Protection Strategies, p. 4-35.

SVO

Caution

High

Ventilation continues as set

Warning

Medium

Ventilation continues as set

Notification

Low

Ventilation continues as set (not displayed on


alarm banner)

Reference the table below for a list of ventilator technical alarms, their meaning, and what to do if they occur.

Operators Manual

6-17

Performance

Reference Alarm Settings Range and Resolution, p. 11-17 for the settings,
ranges, resolutions, new patient default values, and accuracies of all the ventilator alarms.
Table6-4. Technical Alarms
Alarm message

Meaning

What to do

O2 SENSOR

O2 sensor is out of calibration


or has failed.

Re-calibrate or replace O2
sensor.

DEVICE ALERT

Various. Technical alarm category is described. Reference


Technical Alarm Categories, p.
6-17. More information for the
particular technical alarm can
be found in the System diagnostic log, a link to which is
provided on the expanded
alarm banner.

Follow remedy message displayed on GUI.

A non-technical alarm is an alarm caused due to a fault in the patient-ventilator interaction or a fault in the electrical or gas supplies that the practitioner
may be able to alleviate.

Table6-5. Non-technical Alarm Summary

6-18

Base message

Priority

AC POWER LOSS

Low

Analysis
message
Operating on
vent main battery.

Remedy
message
N/A

Comments
Ventilators power
switch is ON. Ventilator automatically
switches to battery
power. AC power
not available.
Battery operating
indicator on status
display turns on.
Resets when AC
power is restored.

Operators Manual

Ventilator Settings

Table6-5. Non-technical Alarm Summary (Continued)


Base message
APNEA (patient
data alarm)

CIRCUIT
DISCONNECT

COMPLIANCE
LIMITED VT
(alarm is not
adjustable)
(patient data
alarm)

Operators Manual

Priority

Analysis
message

Medium

Apnea ventilation. Breath


interval > apnea
interval.

High

Extended apnea
duration or multiple apnea
events.

High

Remedy
message

Comments

Check patient &


settings.

The set apnea interval has elapsed


without the ventilator, patient, or
operator triggering
a breath. Resets
after patient initiates a third consecutive breath.
Possible dependent
alarm: 3VE TOT

No ventilation

Check patient
Reconnect circuit.

Ventilator has
recovered from
unintended power
loss lasting more
than five minutes,
detects circuit disconnect. The GUI
screen displays
elapsed time
without ventilator
support. Resets
when patient is
reconnected.

High

No ventilation

Check patient.
Reconnect circuit.

Ventilator detects
circuit disconnect;
the GUI screen displays elapsed time
without ventilator
support. Resets
when patient is
reconnected.

Low

Compliance
compensation
limit reached

Check patient and


circuit type.
Inspired volume
may be < set.

Compliance volume
required to compensate delivery of
a VC, VC+ or VS
breath exceeds the
maximum allowed
for 3 of the last 4
breaths.

6-19

Performance

Table6-5. Non-technical Alarm Summary (Continued)


Base message
1PPEAK (patient
data alarm)

1PCOMP (patient
data alarm)

Priority

Analysis
message

Low

Last breath set


limit.

Medium

Last 3 breaths
set limit.

High

Last 4 or more
breaths set
limit.

Low

Last spont
breath set
PPEAK limit - 5
cmH2O.

Medium

High

Last 3 spont
breaths set
PPEAK limit - 5
cmH2O.
Last 4 or more
spont breaths
set PPEAK limit- 5
cmH2O.

Remedy
message

Comments

Check patient,
circuit & ET tube.

Measured airway
pressure set limit.
Ventilator truncates current breath
unless already in
exhalation. Possible
dependent alarms:
3VTE MAND,
3 VE TOT, 1fTOT. Corrective action:
Check patient.
Check tube type/ID
setting. Consider
reducing % Supp
setting or increasing 2PPEAK.

In TC:

Pressure of spontaneous breaths set


limit. Possible
dependent alarms:
3 VTE SPONT,
3 VE TOT, 1fTOT,
3VTE SPONT
Corrective action:
Check for leaks.
Check for the
correct tube type.
Check that the tube
inside diameter corresponds to the
patient PBW.

Check for
leaks, tube
type/ID setting.

Check the 2PPEAK


setting.

6-20

Operators Manual

Ventilator Settings

Table6-5. Non-technical Alarm Summary (Continued)


Base message
3PPEAK (patient
data alarm)

Priority

Analysis
message

Low

Last 2 breaths,
pressure set
limit.

Medium

Last 4 breaths,
pressure set
limit

High

Last10 or more
breaths, pressure set limit

Remedy
message
Check for leaks.

Comments
Peak inspiratory
pressure alarm
setting. (Available
only when Mandatory Type is VC+* or
when Vent Type is
NIV. Target pressure = the low limit:
PEEP + 3 cmH2O.
Ventilator cannot
deliver target
volume. Possible
dependent alarms:
1fTOT.
Corrective action:
Check patient and
settings; check for
leaks.

* Because the VC+ pressure control algorithm does not allow the target inspiratory pressure to fall
below PEEP + 3 cmH2O, attempting to set the 4PPEAK alarm limit at or below this level will turn the
alarm off.
1O2% (patient
data alarm)

Operators Manual

Medium

Measured O2%
> set for 30 s
but < 2 min

High

Measured O2%
> set for 2
min.

Check patient, gas


sources, O2 analyzer & ventilator.

The O2% measured


during any phase of
a breath cycle is 7%
(12% during the
first hour of operation) or more above
the O2% setting for
at least 30 seconds.
(These percentages
increase by 5% for
four minutes following a decrease
in the O2% setting.)

6-21

Performance

Table6-5. Non-technical Alarm Summary (Continued)


Base message

Priority

Analysis
message

Remedy
message

Comments

3O2% (patient
data alarm)

High

Measured O2%
< set O2%.

Check patient, gas


sources, O2 analyzer & ventilator.

The O2% measured


during any phase of
a breath cycle is 7%
(12% during the
first hour of operation) or more below
the O2% setting for
at least 30 seconds,
or below 18%.
(These percentages
increase by 5% for
four minutes following an increase
in the O2% setting.)

1VTE (patient
data alarm

Low

Last 2 breaths
set limit.

Medium

Last 4 breath s
set limit.

Check settings,
changes in
patient's R&C.

High

Last 10 or more
breaths set
limit.

Exhaled tidal
volume set limit.
Alarm updated
whenever exhaled
tidal volume is
recalculated. Possible dependent
alarm: 1VE TOT

Low

VE TOT set limit


for 30 s

Check patient and


settings.

Medium

VE TOT set limit


for > 30 s

High

VE TOT set limit


for > 120 s

Expiratory minute
volume set limit.
Alarm updated
whenever an
exhaled minute
volume is recalculated. Possible
dependent alarm:
1VTE.

Low

fTOT set limit


for 30 s.

Check patient &


settings.

Medium

fTOT set limit


for > 30 s.

High

fTOT set limit


for > 120 s.

Total respiratory
rate set limit.
Alarm updated at
the beginning of
each inspiration.
Reset when measured respiratory
rate falls below the
alarm limit. Possible
dependent alarms:
3VTE MAND,
3VTE SPONT, 3VE TOT.

1VE TOT (patient


data alarm)

1fTOT (patient
data alarm)

6-22

Operators Manual

Ventilator Settings

Table6-5. Non-technical Alarm Summary (Continued)


Base message
1PVENT (patient
data alarm)

Priority

Analysis
message

Low

1 breath limit.

Medium

2 breaths limit.

High

INOPERATIVE
BATTERY

Operators Manual

Low

Remedy
message
Check patient,
circuit & ET tube.

3 or more
breaths limit.

Inadequate
charge or nonfunctional ventilators primary
battery.

Service/replace
ventilators primary
battery.

Comments
Inspiratory pressure
> 100 cmH2O and
mandatory type is
VC or spontaneous
type is TC or PAV+.
Ventilator truncates current breath
unless already in
exhalation. Possible
dependent alarms:
3VTE MAND,
3VE TOT, 1fTOT.
Corrective action:
1.

Check patient
for agitation.
Agitated
breathing,
combined with
high % Supp
setting in PAV+
can cause
over-assistance. Consider reducing %
Supp setting.

2.

Provide alternate ventilation. Remove


ventilator from
use and
contact Service.

Battery installed but


not functioning or
charging for 6
hours. Resets when
battery is functional.

6-23

Performance

Table6-5. Non-technical Alarm Summary (Continued)


Base message
INSPIRATION
TOO LONG
(patient data
alarm)

PAV STARTUP
TOO LONG
(patient data
alarm) (occurs
only if PAV+ is in
use)

6-24

Priority

Analysis
message

Low

Last 2 spont
breaths = PBW
based TI limit.

Medium

Last 4 spont
breaths = PBW
based TI limit.

High

Last 10 or more
spont breaths =
PBW based TI
limit.

Low

PAV startup not


complete for
45 s.

Medium

PAV startup not


complete for
90 s.

High

PAV startup not


complete for
120 s.

Remedy
message

Comments

Check patient.
Check for leaks.

Inspiratory time for


spontaneous breath
PBW-based limit.
Ventilator transitions to exhalation.
Resets when TI falls
below PBW-based
limit. Active only
when Vent Type is
INVASIVE.

Check for leaks,


shallow breathing,
& settings for
1VTI and 1PPEAK.

Unable to assess
patients resistance
and compliance
during PAV startup.
Possible dependent
alarms 3VTE SPONT,
3VE TOT, 1fTOT. Corrective action:
Check patient.
(Patients inspiratory times may be too
short to evaluate
resistance and compliance.) Check that
selected humidification type and empty
humidifier volume
are correct.

Operators Manual

Ventilator Settings

Table6-5. Non-technical Alarm Summary (Continued)


Base message

Priority

Analysis
message

Remedy
message

Comments

PAV R & C NOT


ASSESSED
(patient data
alarm) (occurs
only if PAV+ is in
use)

Low

R and/or C over
15 minutes old.

Medium

R and/or C over
30 minutes old.

Check for leaks,


shallow breathing,
& settings for tube
ID, 1VTI and
1PPEAK.

Unable to assess
resistance and/or
compliance during
PAV+ steady-state.
Startup was successful, but later
assessments were
unsuccessful. Corrective action:
Check patient.
(Patients inspiratory times may be too
short to evaluate
resistance and compliance.) Check that
selected humidification type and empty
humidifier volume
are correct.

LOSS OF POWER

Immediate

N/A

N/A

The ventilator
power switch is ON
and there is insufficient power from
AC and the battery.
There may not be a
visual indicator for
this alarm, but an
independent audio
alarm sounds for at
least 120 seconds.
Alarm annunciation can be reset by
turning power
switch to OFF position.

Operating on
battery.Ventilator battery operational time <
10 minutes.

Replace or allow
ventilator battery
to recharge.

Resets when
battery has ten
minutes of operational time remaining or when AC
power is restored.

LOW BATTERY

Operators Manual

Medium

6-25

Performance

Table6-5. Non-technical Alarm Summary (Continued)


Base message

Priority

Analysis
message

LOW BATTERY

High

Operating on
battery. Ventilator battery operational time < 5
minutes

Replace or allow
ventilator battery
to recharge

Resets when
battery has five
minutes of operational time remaining or when AC
power is restored.

3VTE MAND
(patient data
alarm)

Low

Last 2 mand
breaths set
limit.

Check for leaks,


changes in
patient's R & C.

Medium

Last 4 mand
breaths set
limit.

Exhaled mandatory
tidal volume set
limit. Alarm
updated whenever
exhaled mandatory
tidal volume is
recalculated. Possible dependent
alarms: 3VE TOT,
1fTOT.

Check patient &


settings.

Exhaled spontaneous tidal volume


set limit. Alarm
updated whenever
exhaled spontaneous tidal volume
is recalculated. Possible dependent
alarms: 3VE TOT,
1fTOT.

3VTE SPONT
(patient data
alarm)

High

Last 10 or more
mand breaths
set limit.

Low

Last 4 spont
breaths set
limit

Medium

Last 7 spont
breaths set
limit

High

6-26

Last 10 or more
spont breaths
set limit

Remedy
message

Comments

Operators Manual

Ventilator Settings

Table6-5. Non-technical Alarm Summary (Continued)


Base message

Priority

1VTI (patient data


alarm)

Low

Last spont
breath set
limit.

In TC, VS, or
PAV+:

Medium

Last 3 spont
breaths set
limit.

High

3VE TOT (patient


data alarm)

Operators Manual

Analysis
message

Remedy
message

Check patient
and settings.

Last 4 or more
spont breaths
set limit.

Low

VE TOT set limit


for 30 s.

Medium

VE TOT set limit


for > 30 s.

High

VE TOT set limit


for > 120 s.

Comments
Delivered inspiratory volume inspiratory limit. Ventilator
transitions to exhalation. Possible
dependent alarms:
3VTE SPONT,3VE TOT,
1fTOT
Corrective action:
Check for leaks.
Check for the
correct tube type.
Check the VTI or VTI
setting. In PAV+,
check for patient
agitation, which
can cause miscalculation of RPAV and
CPAV. Consider
reducing % Supp
setting. Check 2VTI.

Check patient &


settings.

Total minute
volume set limit.
Alarm updated
whenever exhaled
minute volume is
recalculated. Possible dependent
alarms3VTE MAND,
3VTE SPONT, 1fTOT

6-27

Performance

Table6-5. Non-technical Alarm Summary (Continued)


Base message
VOLUME NOT
DELIVERED (not
adjustable)
(patient data
alarm)

Priority

Analysis
message

Remedy
message

Comments

Low

Last 2 spont (or


mand) breaths
pressure > max
allowable level.

Check patient and


setting for 1PPEAK.

Medium

Last 10 or more
spont (or mand)
breaths, pressure > max
allowable level.

Insp target pressure


> (PPEAK - PEEP - 3
cmH2O), when
spontaneous type is
VS or mandatory
type is VC+. Ventilator cannot deliver
target volume. Possible dependent
alarms: For VC+
breaths: 3VTE MAND
3VE TOT 1fTOT For VS
breaths: 3VTE

SPONT,

1fTOT

3VE TOT,

Corrective action:
Check patient and
settings.
NO AIR SUPPLY

High

Ventilation continues as set


except O2=
100%

Check patient and


air source

Switch to 100% O2.

NO O2 SUPPLY

Low

Ventilation continues as set.


Only air available.

Check O2 source.

Operator-set O2%
equals 21%. Resets
if O2 supply connected

High

Ventilation continues as set,


except O2% =
21.

Check patient & O2


source.

Ventilator delivers
21% O2 instead of
set O2%. Resets if
oxygen supply connected.

High

Patient connected before setup


complete

Provide alternate
ventilation. Complete setup process.

Ventilator begins
safety ventilation.
Resets when ventilator startup procedure is complete.

PROCEDURE
ERROR

6-28

Operators Manual

Ventilator Settings

Table6-5. Non-technical Alarm Summary (Continued)


Base message

Priority

Analysis
message

SEVERE
OCCLUSION

High

Little/no ventilation.

Check patient.
Provide alternate
ventilation. Clear
occlusions; drain
circuit.

Ventilator enters
occlusion status
cycling (OSC).
Patient data displays are blanked
and GUI screen displays elapsed time
without ventilator
support.

PROX INOPERATIVE (if Proximal


Flow Sensor is in
use)

Low

Data from Proximal Flow Sensor


are not being
used.

Check proximal
flow sensor connections and tubes
for occlusions or
leaks

Data obtained from


the proximal flow
sensor are invalid,
non-existent, or
unreasonable based
on current ventilator settings or
purge lines are
occluded. Alarm
resets when condition is corrected.
Data for real time
waveforms and
monitored volumes
are obtained from
internal sensors.

Ventilator
switched OFF
with patient
connected to
breathing circuit.

Return power
switch to ON position and disconnect patient before
turning power off.

User must acknowledge turning the


power OFF by
touching Power Off
on the GUI.

INADVERTENT
POWER OFF

High

Remedy
message

Comments

Table6-6.Non-Technical Alarms and Suggested Responses


Alarm message

Meaning

Response

AC POWER LOSS

The ventilator is running on


battery power.

Monitor the battery charge


level to ensure there is enough
power remaining to operate
the ventilator.

APNEA (patient data alarm)

The time between patient


breaths exceeds the set apnea
interval.

Check patient and settings.

Operators Manual

6-29

Performance

Table6-6.Non-Technical Alarms and Suggested Responses (Continued)


Alarm message

Meaning

Response

CIRCUIT DISCONNECT

The patient circuit has become


disconnected or there is a large
leak in the patient circuit.

Re-connect the patient circuit,


or eliminate the leak.

Compliance limited VT (patient


data alarm)

Compliance volume required to


compensate delivery of a VC,
VC+, or VS breath exceeds the
maximum allowed for three of
the last four breaths.

Check patient and circuit type.


Inspired volume may be less
than set.

1PPEAK (patient data alarm)

The measured airway pressure


is set limit. Reduced tidal
volume likely.

Check the patient

Check the patient circuit

Check the endotracheal


tube

3PPEAK (patient data alarm)

The peak inspiratory pressure in


the patient circuit alarm setting.This alarm is only available
when NIV is the selected Vent
Type or when VC+ is the selected Mandatory type during
INVASIVE ventilation.*

Check for leaks in the patient


circuit and VBS.

* Because the VC+ pressure control algorithm does not allow the target inspiratory pressure to fall
below PEEP + 3 cmH2O, attempting to set the 4PPEAK alarm setting at or below this level will turn the
alarm off.
1O2% (patient data alarm)

6-30

The O2% measured during any


phase of a breath cycle is 7%
(12% during the first hour of
operation) or more above the
O2% parameter for at least 30
seconds.The percentage
window increases by 5% for
four minutes after increasing
the set O2% value.

Check the patient, the air and


oxygen supplies, the oxygen
analyzer, and the ventilator.

Operators Manual

Ventilator Settings

Table6-6.Non-Technical Alarms and Suggested Responses (Continued)


Alarm message
3O2% (patient data alarm)

1VTE (patient data alarm)

Meaning
The O2% measured during any
phase of a breath cycle is 7%
(12% during the first hour of
operation) or more below the
O2% parameter for at least 30
seconds. The percentage
window increases by 5% for
four minutes after increasing
the set O2% value.

Exhaled tidal volume alarm


setting for the last two breaths.

Response

Check the patient, the air


and oxygen supplies, the
oxygen analyzer, and the
ventilator.

Calibrate oxygen sensor.


Reference Oxygen Sensor
Calibration, p. 4-34 for
details regarding calibrating the oxygen sensor.

Use an external O2
monitor and disable the O2
sensor.

Check patient settings

Check for changes in


patients resistance and
compliance.

1VE TOT (patient data alarm)

Minute volume alarm setting.

Check patient settings.

1fTOT (patient data alarm)

The breath rate from all breaths


is alarm setting.

Check the patient and the ventilator settings.

1PVENT (patient data alarm)

The inspiratory pressure transducer has measured a pressure


> 110 cmH2O in VC, TC, or
PAV+. The ventilator transitions
to exhalation. A reduced tidal
volume is likely.

Check the patient, the


patient circuit (including
filters), and the endotracheal tube. Ensure the ET
tube ID is the correct size.
Check the ventilator flow
and/or volume settings.

Re-run SST.

Obtain an alternate ventilation source.

Remove the ventilator


from clinical use and
obtain service.

INOPERATIVE BATTERY

Operators Manual

The battery charge is inadequate after 6 hours of attempted charge time or the battery
system is non-functional.

Replace the battery or install an


extended battery.

6-31

Performance

Table6-6.Non-Technical Alarms and Suggested Responses (Continued)


Alarm message
INSPIRATION TOO LONG
(patient data alarm)

LOSS OF POWER

Meaning
The PBW-based inspiratory time
for the last two spontaneous
breath exceeds the ventilatorset limit. Active only when Vent
Type is INVASIVE.

The ventilator power switch is


ON, but there is insufficient
power from the mains AC and
the battery. There may not be a
visual indicator for this alarm,
but an independent audio
alarm (immediate priority)
sounds for at least 120 seconds.

Response

Check the patient.

Check the patient circuit


for leaks.

Check Rise time % and


ESENS settings.

Check the integrity of the


AC power and battery
connections.

Obtain alternative ventilation if necessary.

Install an extended battery

If the loss of power event has


been resolved, turn the
power switch OFF and
back ON again to reset the
alarm.
LOW BATTERY

Medium priority alarm indicating < ten (10) minutes of


battery power remaining to
operate the ventilator.
High priority alarm indicating <
five (5) minutes of battery
power remain to operate the
ventilator.

Recharge the battery, by plugging the ventilator into AC


power or replace the battery, or
install an extended battery.

3VTE MAND (patient data alarm)

The patients exhaled mandatory tidal volume is alarm


setting for the last two mandatory breaths.

Check the patient.

Check for leaks in the


patient circuit.

Check for changes in the


patients resistance or
compliance.

Check the patient.

Check the ventilator settings.

3VTE SPONT (patient data alarm)

6-32

The patients exhaled spontaneous tidal volume is alarm


setting for the last two spontaneous breaths.

Operators Manual

Ventilator Settings

Table6-6.Non-Technical Alarms and Suggested Responses (Continued)


Alarm message
3VE TOT (patient data alarm)

NO AIR SUPPLY

NO O2 SUPPLY

1PCOMP

Meaning
The minute volume for all
breaths is alarm setting.

Response

Check the patient.

Check the ventilator settings.

The air supply pressure is less


than the minimum pressure
required for correct ventilator
operation. The ventilator delivers 100% O2 if available. If an
oxygen supply is not available,
the safety valve opens. The ventilator displays the elapsed time
without ventilator support. This
alarm cannot be set or disabled.

Check the patient.

Check the air and oxygen


sources.

Obtain alternative ventilation if necessary.

The oxygen supply pressure is


less than the minimum pressure
required for correct ventilator
operation. The ventilator delivers 100% air if available. If an
air supply is not available, the
safety valve opens. The ventilator displays the elapsed time
without ventilatory support.
This alarm cannot be set or disabled.

Check the patient.

Check the air and oxygen


sources.

Obtain alternative ventilation if necessary.

Target pressure (2PPEAK - 5


cmH2O

In TC:

Check for leaks and tube


type/I.D. setting

In PAV+:

PROCEDURE ERROR

Operators Manual

The patient is attached before


ventilator startup is complete.
Safety ventilation is active.

Limit target pressure to


2PPEAK - 5 cmH2O

Provide alternate ventilation if necessary.

Complete ventilator
startup procedure.

6-33

Performance

Table6-6.Non-Technical Alarms and Suggested Responses (Continued)


Alarm message
SEVERE OCCLUSION

1VTI (patient data alarm)

Meaning
The patient circuit is severely
occluded. The ventilator enters
occlusion status cycling. The
elapsed time without ventilatory support appears.

Delivered inspiratory volume


high inspiratory volume limit

Response

Check the patient.

Obtain alternative ventilation if necessary.

Check patient circuit for


bulk liquid, crimps,
blocked filter.

If problem persists, remove


ventilator from use and
obtain service.

Ventilator transitions to exhalation.

Check for leaks and tube


type/ID setting.

Check patient and ventilatorsettings.

Check for leaks, tube type/


ID and% Supp settings, and
patient agitation.

VOLUME NOT DELIVERED


(patient data alarm

Insp target pressure > (PPEAK PEEP - 3 cmH2O), when spontaneous type is VS or when mandatory type is VC+.

Check patient and 1PPEAK


setting

PAV STARTUP TOO LONG


(occurs only if PAV+ option is in
use)

Unable to assess resistance and/


or compliance during PAV+
startup.

Check for leaks, shallow


breathing, and settings for
1VTI and 1PPEAK

PAV R & C NOT ASSESSED


(occurs only if PAV+ option is in
use)

Unable to assess resistance and/


or compliance during PAV+
steady-state.

Check for leaks, shallow


breathing, and settings for tube
ID, 1VTI and 1PPEAK.

PROX INOPERATIVE

A malfunction occurred with


the Proximal Flow Sensor or the
pneumatic lines are occluded.

Replace the Proximal Flow


Sensor or purge its pneumatic
lines. Does not affect data from
the ventilators delivery or exhalation valve flow sensors.

The next sections provide detailed descriptions of selected alarms.

6-34

Operators Manual

Ventilator Settings

6.5.9

AC POWER LOSS Alarm


The AC POWER LOSS alarm indicates the ventilator power switch is on and the
ventilator is being powered by the battery and an alternate power source may
soon be required to sustain normal ventilator operation. The ventilator annunciates a medium-priority LOW BATTERY alarm when the ventilator has less
than ten minutes of battery power remaining. The ventilator annunciates a
high-priority LOW BATTERY alarm when less than five minutes of battery
power are estimated available.

6.5.10

Apnea Alarm
The APNEA alarm indicates neither the ventilator nor the patient has triggered
a breath for the operator-selected apnea interval (TA). TA is measured from the
start of an inspiration to the start of the next inspiration and is based on the
ventilators inspiratory detection criteria. TA can only be set via the apnea ventilation settings.
The APNEA alarm autoresets after the patient initiates two successive breaths,
and is intended to establish the patient's inspiratory drive is reliable enough to
resume normal ventilation. To ensure the breaths are patient-initiated (and not
due to autotriggering), exhaled volumes must be at least half the inspired VT
(this avoids returning to normal ventilation if there is a disconnect).

6.5.11

CIRCUIT DISCONNECT Alarm


The CIRCUIT DISCONNECT alarm indicates the patient circuit is disconnected
at the ventilator or the patient side of the patient wye, or a large leak is present. The methods by which circuit disconnects are detected vary depending on
breath type. Time, pressure, flow, delivered volume, exhaled volume, and the
DSENS setting may be used in the circuit disconnect detection algorithms. Reference Disconnect, p. 10-47 for a complete discussion of the CIRCUIT DISCONNECT detection methods.
The CIRCUIT DISCONNECT alarm sensitivity is adjusted via the DSENS setting.
During a CIRCUIT DISCONNECT condition, the ventilator enters an idle state
and delivers a base flow of oxygen to detect a reconnection.
When the ventilator determines the patient circuit is reconnected, the CIRCUIT
DISCONNECT alarm autoresets and normal ventilation resumes without having
to manually reset the alarm (for example, following suctioning).

Operators Manual

6-35

Performance

A disconnected patient circuit interrupts gas delivery and patient monitoring.


Notification of a patient circuit disconnect is crucial, particularly when the
patient cannot breathe spontaneously. The ventilator does not enter apnea
ventilation when a disconnect is detected to avoid changing modes during a
routine suctioning procedure.

6.5.12

Note:

When utilizing a closed-suction catheter system, the suctioning procedure can


be executed using existing mode, breath type and settings. To reduce potential
for hypoxemia during the procedure, elevate the oxygen concentration using
the Elevate O2 control. Reference To adjust the amount of elevated O2 delivered for
two minutes, p. 3-40.

Loss of Power Alarm


This alarm alerts the operator that there is insufficient battery power and no
AC power to support ventilator operation. The alarm annunciates as long as
the ventilators power switch is in the ON position, and lasts for at least 120
seconds.

6.5.13

DEVICE ALERT Alarm


A DEVICE ALERT alarm indicates a background test or Power On Self Test
(POST) has failed. Depending on which test failed, the ventilator either continues to ventilate according to current settings, or ventilates with modified settings, or enters the ventilator inoperative state. The DEVICE ALERT alarm relies
on the ventilators self-testing and notifies the clinician of an abnormal condition requiring service. Reference Background Diagnostic System, p. 10-74.

6.5.14

High Circuit Pressure (1PPEAK) Alarm


The 1PPEAK alarm indicates the currently measured airway pressure is equal to
or greater than the set limit. The 2PPEAK limit is active during all breath types
and phases to provide redundant patient protection (for example, to detect air
flow restrictions downstream of the pressure-sensing device). The 1PPEAK limit
is active in all normal ventilation modes. The 2 PPEAK alarm new patient default
values are separately configurable for neonatal, pediatric, and adult patients.
The2PPEAK limit is not active during a SEVERE OCCLUSION alarm.

6-36

Operators Manual

Ventilator Settings

The 1PPEAK alarm truncates inspiration and transitions the ventilator into the
exhalation phase and the limit cannot be set less than

PEEP + 7 cmH2O, or

PEEP + PI + 2 cmH2O, or

PEEP + PSUPP + 2 cmH2O

nor can it be set less than or equal to 4PPEAK.


The 2PPEAK limit cannot be disabled. The ventilator phases in changes to the
2PPEAK limit immediately to allow prompt notification of a high circuit pressure
condition.
The minimum 2PPEAK limit (7 cmH2O) corresponds to the lowest peak pressures not due to autotriggering anticipated during a mandatory breath. The
maximum 2PPEAK limit (100 cmH2O) was selected because it is the maximum
pressure required to inflate very low-compliance lungs.
The ventilator allows circuit pressure to rise according to a computed triggering
profile for the initial phase of PC and PS breaths without activating the 2PPEAK
alarm. This triggering profile helps avoid nuisance alarms due to possible transient pressure overshoot in the airway when aggressive values of rise time %
are selected. A brief pressure overshoot measured in the patient circuit is
unlikely to be present at the carina.
6.5.15

High Delivered O2% (1O2%) Alarm


The 1O2% alarm indicates the measured O2% is at or above the error percentage above the O2% setting for at least 30 seconds to eliminate transient O2%
delivery variation nuisance alarms. The 1O2% alarm detects malfunctions in
ventilator gas delivery or oxygen monitor. The ventilator declares a 1O2%
alarm after 30 seconds. Although the ventilator automatically sets the 1O2%
alarm limits, the oxygen sensor can be disabled. (The error percentage is 12%
above setting for the first hour of ventilator operation, 7% above the setting
after the first hour of operation, and an additional 5% above the setting for
the first four minutes following a decrease in the setting.)
The ventilator automatically adjusts the 1O2% alarm limit when O2% changes
due to 100% O2, apnea ventilation, occlusion, circuit disconnect, or a NO AIR/

Operators Manual

6-37

Performance

O2 SUPPLY alarm. The ventilator checks the 1O2% alarm limit against the measured oxygen percentage at one-second intervals.
6.5.16

High Exhaled Minute Volume (1VE TOT) Alarm


The 1VE TOT alarm indicates the measured exhaled total minute volume for
spontaneous and mandatory breaths is equal to or greater than the alarm setting. The 1VE TOT alarm is effective immediately upon changing the setting, to
ensure prompt notification of prolonged high tidal volumes.
The 1VE TOT alarm can be used to detect a change in a patient's breathing pattern, or a change in compliance or resistance. The 1VE TOT alarm can also detect
too-large tidal volumes, which could lead to hyperventilation and hypocarbia.

6.5.17

High Exhaled Tidal Volume (1VTE) Alarm


The 1VTE alarm indicates the measured exhaled tidal volume for spontaneous
and mandatory breaths is equal to or greater than the set 1VTE alarm. The 1VTE
alarm is updated whenever a new measured value is available.
The 1VTE alarm can detect increased exhaled tidal volume (due to greater compliance and lower resistance) and prevent hyperventilation during pressure
control ventilation or pressure support. Turn the 1VTE alarm OFF to avoid nuisance alarms. (Hyperventilation due to increased compliance is not a concern
during volume-based ventilation, because the tidal volume is fixed by the clinician's choice and the ventilators compliance-compensation algorithm.)

6.5.18

High Inspired Tidal Volume (1VTI) Alarm


The high inspired tidal volume alarm indicates the patients inspired volume
exceeds the set limit. When this condition occurs, the breath terminates and
the alarm sounds. The ventilator displays monitored inspired tidal volume
values in the patient data area on the GUI screen. When Vent Type is NIV, there
is no high inspired tidal volume alarm or setting available, but the monitored
inspired tidal volume (VTI) may appear in the patient data area on the GUI
screen.

6-38

Operators Manual

Ventilator Settings

6.5.19

High Respiratory Rate 1(fTOT) Alarm


The 1fTOT alarm indicates the measured breath rate is greater than or equal to
the 2fTOT alarm setting. The 1fTOT alarm is updated whenever a new total measured respiratory rate is available.
The 1fTOT alarm can detect tachypnea, which could indicate the tidal volume
is too low or the patient's work of breathing has increased. The ventilator
phases in changes to the 2fTOT limit immediately to ensure prompt notification
of a high respiratory rate condition.

6.5.20

INSPIRATION TOO LONG Alarm


The INSPIRATION TOO LONG alarm, active only when Vent Type is INVASIVE,
indicates the inspiratory time of a spontaneous breath exceeds the following
time limit
(1.99 + 0.02 x PBW) seconds (adult and pediatric circuits)
(1.00 + 0.10 x PBW) seconds (neonatal circuits)
where PBW is the current setting for predicted body weight in kg.
When the ventilator declares an INSPIRATION TOO LONG alarm, the ventilator
terminates inspiration and transitions to exhalation. The INSPIRATION TOO
LONG alarm applies only to spontaneous breaths and cannot be set or disabled.
Because leaks (in the patient circuit, around the endotracheal tube cuff, or
through chest tubes) and patient-ventilator mismatch can affect accurate
exhalation detection, the INSPIRATION TOO LONG alarm can act as a backup
method of safely terminating inspiration. If the INSPIRATION TOO LONG alarm
occurs frequently, check for leaks and ensure ESENS and rise time % are properly set.

6.5.21

Low Circuit Pressure (3PPEAK) Alarm


The 3PPEAK alarm indicates the measured maximum airway pressure during the
current breath is less than or equal to the set alarm level during a non-invasive
inspiration or during a VC+ inspiration.

Operators Manual

6-39

Performance

The 3PPEAK alarm is active for mandatory and spontaneous breaths, and is
present only when Vent Type is NIV or Mandatory Type is VC+. During VC+,
the 3PPEAK alarm can be turned OFF. The 3PPEAK alarm can always be turned
OFF during NIV. The 4PPEAK alarm limit cannot be set to a value greater than
or equal to the 2PPEAK alarm limit.

WARNING:
Because the VC+ pressure control algorithm does not allow the target
inspiratory pressure to fall below PEEP + 3 cmH2O, attempting to set the
4PPEAK alarm limit at or below this level will turn the alarm off.

In VC+, whenever PEEP is changed, 3PPEAK is set automatically to its New


Patient value, PEEP + 4 cmH2O when PEEP 16 cmH2O, or PEEP + 3.5 cmH2O
when PEEP < 16 cmH2O.
There are no alarms dependent upon 3PPEAK, and the 3PPEAK alarm does not
depend on other alarms.
6.5.22

Low Delivered O2% (3O2%) Alarm


The 3O2% alarm indicates the measured O2% during any phase of a breath is
at or below the error percentage below the O2% setting, or less than or equal
to 18%, for at least 30 seconds. Although the ventilator automatically sets the
3O2% alarm, replace (if necessary) or disable the oxygen sensor to avoid nuisance alarms. (The error percentage is 12% below setting for the first hour of
ventilator operation following a reset, 7% below setting after the first hour of
operation, and an additional 5% below setting for the first four minutes following an increase in the setting.)
The ventilator automatically adjusts the 3O2% alarm limit when O2% changes
due to apnea ventilation, circuit disconnect, or a NO O2/AIR SUPPLY alarm. The
3O2% alarm is disabled during a safety valve open (SVO) condition. The ventilator checks the 3O2% alarm against the measured oxygen percentage at onesecond intervals.
The 3O2% alarm can detect malfunctions in ventilator gas delivery or the
oxygen monitor, and can ensure the patient is adequately oxygenated. The
ventilator declares a 3O2% alarm after 30 seconds to eliminate nuisance
alarms from transient O2% delivery variations. The O2% measured by the

6-40

Operators Manual

Ventilator Settings

oxygen sensor is shown in the patient data area. Reference Vital Patient Data,
p. 3-42 to include O2% if it is not displayed.
6.5.23

Low Exhaled Mandatory Tidal Volume (3VTE MAND) Alarm


The alarm indicates the measured exhaled mandatory tidal volume is less than
or equal to the 3VTE MAND alarm setting. The 3VTE MAND alarm updates when
a new measured value of exhaled mandatory tidal volume is available.
The 3VTE MAND alarm can detect an obstruction, a leak during volume ventilation, or a change in compliance or resistance during pressure-based ventilation
(that is, when the same pressure is achieved but tidal volume decreases). There
are separate alarms for mandatory and spontaneous exhaled tidal volumes for
use during SIMV, SPONT, and BiLevel. The ventilator phases in a change to the
3VTE MAND alarm immediately to ensure prompt notification of a low exhaled
tidal volume condition.

6.5.24

Low Exhaled Spontaneous Tidal Volume (3VTE SPONT) Alarm


The 3VTE SPONT alarm indicates the measured exhaled spontaneous tidal
volume is less than or equal to the 3VTE SPONT alarm setting. The alarm updates
when a new measured value of exhaled spontaneous tidal volume is available.
The 3VTE SPONT alarm can detect a leak in the patient circuit or a change in the
patients respiratory drive during a single breath. The 3VTE SPONT alarm is based
on the current breath rather than on an average to detect changes as quickly
as possible. There are separate alarms for mandatory and spontaneous exhaled
tidal volumes for use during SIMV and BiLevel (if this software option is
installed). The ventilator phases in a change to the 4VTE SPONT alarm limit
immediately to ensure prompt notification of a low exhaled tidal volume condition.

6.5.25

Low Exhaled Total Minute Volume (3VE TOT) Alarm


The 3VE TOT alarm indicates the measured minute volume (for mandatory and
spontaneous breaths) is less than or equal to the 3VE TOT alarm setting. The
3VE TOT alarm updates with each new calculation for exhaled minute volume.

Operators Manual

6-41

Performance

The 3VE TOT alarm can detect a leak or obstruction in the patient circuit, a
change in compliance or resistance, or a change in the patient's breathing pattern. The 3VE TOT alarm can also detect too-small tidal volumes, which could
lead to hypoventilation and hypoxia (oxygen desaturation).
The ventilator phases in changes to the 3VE TOT alarm limit immediately to
ensure prompt notification of prolonged low tidal volumes.
6.5.26

PROCEDURE ERROR Alarm


The ventilator declares a PROCEDURE ERROR alarm if it is powered up (either
by turning on the power switch or if power is regained following a power loss
of at least five minutes) and the ventilator detects a patient attached before
Ventilator Startup is complete. Until confirmation of the ventilator settings, the
ventilator annunciates a high-priority alarm and enters Safety PCV. Reference
Safety PCV Settings, p. 10-72.
The PROCEDURE ERROR alarm requires confirmation of ventilator settings
after restoration of ventilator power, in case a new patient is attached to the
ventilator. Safety PCV is an emergency mode of ventilation providing ventilation according to displayed settings until settings confirmation, and is not
intended for long-term patient ventilation.

6.5.27

Severe Occlusion Alarm


A severe occlusion alarm occurs when gas flow in the ventilator breathing
system is severely restricted. The ventilator enters Occlusion Status Cycling
(OSC) where the ventilator periodically attempts to deliver a pressure-based
breath while monitoring inspiration and exhalation breath phases for a severe
occlusion. If an occlusion is not detected, the ventilator considers the occlusion
condition reset, clears the occlusion alarm, and continues ventilation with the
settings in use before the occlusion occurred. The ventilator indicates an occlusion was detected.

6.6

Monitored Patient Data


Monitored patient data appear in the Patient Data Banner at the top of the GUI
screen above the waveforms display. Reference Areas of the GUI, p. 4-3.

6-42

Operators Manual

Ventilator Settings

Reference Vital Patient Data, p. 3-42 to change the displayed patient data
parameters or the order in which they are displayed.
If any patient data values are displayed continuously blinking, it means their
values are shown clipped to what has been defined as their absolute limits. If
the values are displayed in parentheses ( ), it means they are clipped to their
variable limits.Variable limits are calculated values derived from the set PBW
and ventilator settings. Displayed patient data values that have been clipped
should be viewed as suspect.
Dashes (--) are displayed if the patient data value is not applicable based on
mode/breath type combinations.

Note:
If no value is displayed, then the ventilator is in a state where the value cannot be
measured.

The following sections contain descriptions of all patient data parameters


shown in the patient data displays.

6.6.1

Note:
All displayed patient volume data represent lung volumes expressed under BTPS
conditions.

Total Exhaled Minute Volume (VE TOT)


The BTPS and compliance compensated sum of exhaled gas volumes from
both mandatory and spontaneous breaths for the previous one-minute interval.

6.6.2

Exhaled Spontaneous Minute Volume (VE SPONT)


The BTPS- and compliance-compensated sum of exhaled spontaneous volumes
for the previous minute.

6.6.3

Exhaled Tidal Volume (VTE)


The volume of the patients exhaled gas for the previous mandatory or spontaneous breath. Displayed VTE is both compliance-and BTPS compensated, and
updates at the next inspiration.

Operators Manual

6-43

Performance

6.6.4

Proximal Exhaled Minute Volume (VE TOTY)


The BTPS- and compliance-compensated sum of exhaled spontaneous volumes
for the previous minute.

6.6.5

Proximal Exhaled Tidal Volume (VTEY)


The exhaled tidal volume for the previous breath measured by the Proximal
Flow Sensor (for neonatal patients, only). VTEY is updated at the beginning of
the next inspiration.

6.6.6

Exhaled Spontaneous Tidal Volume (VTE SPONT)


The exhaled volume of the last spontaneous breath, updated at the beginning
of the next inspiration following a spontaneous breath.

6.6.7

Exhaled Mandatory Tidal Volume (VTE MAND)


The exhaled volume of the last mandatory breath, updated at the beginning
of the next inspiration following a mandatory breath. If the mode is SPONT and
the ventilator has not delivered mandatory breaths in a time period of greater
than two minutes (for example via a manual inspiration), the VTE MAND patient
data indicator becomes hidden.The indicator reappears when the value
updates.

6.6.8

Exhaled mL/kg Volume


The patients exhaled volume displayed in mL/kg PBW.

6.6.9

Inspired Tidal Volume (VTI)


The BTPS- and compliance-compensated volume of inspired gas for all pressure-based or NIV breaths, updated at the beginning of the following expiratory phase. VTI is displayed when data are available.

6-44

Operators Manual

Ventilator Settings

6.6.10

Proximal Inspired Tidal Volume (VTIY)


The inspired tidal volume for a mandatory or spontaneous breath measured by
the Proximal Flow Sensor (for neonatal patients, only). VTIY is updated at the
beginning of the following expiratory phase and is displayed when data are
available.

6.6.11

Delivered mL/kg Volume


The delivered gas volume in mL/kg PBW.

6.6.12

I:E Ratio
The ratio of inspiratory time to expiratory time for the previous breath, regardless of breath type. Updated at the beginning of the next inspiration. When I:E
ratio is 1:1, it is displayed as XX:1. Otherwise it is displayed as 1:XX.

6.6.13

Note:
Due to limitations in setting the I:E ratio in PC ventilation, the monitored data display
may not exactly match the I:E ratio setting.

Mean Circuit Pressure (PMEAN)


The average circuit pressure for a complete breath period, including both inspiratory and expiratory phases whether mandatory or spontaneous. The displayed value can be either positive or negative.

6.6.14

Peak Circuit Pressure (PPEAK)


The maximum circuit pressure at the patient wye during the previous breath,
including both inspiratory and expiratory phases.

6.6.15

End Inspiratory Pressure (PI END)


The pressure at the end of the inspiratory phase of the current breath.

Operators Manual

6-45

Performance

6.6.16

End Expiratory Pressure (PEEP)


The pressure at the end of the expiratory phase of the previous breath,
updated at the beginning of the next inspiration. During an expiratory pause,
the displayed value includes any active lung PEEP.

6.6.17

Intrinsic PEEP (PEEPI)


PEEPI is an estimate of the pressure above the PEEP level at the end of an exhalation. PEEPI is determined during an Expiratory Pause maneuver.

6.6.18

PAV-based Intrinsic PEEP (PEEPI PAV)


PEEPI is an estimate of intrinsic PEEP, updated at the end of a spontaneous
PAV+ breath.

6.6.19

Total PEEP (PEEPTOT)


PEEPTOT is the estimated pressure at the circuit wye during the Expiratory Pause
maneuver.

6.6.20

Plateau Pressure (PPL)


PPL is the pressure measured and displayed during an Inspiratory Pause maneuver.

6.6.21

Total Respiratory Rate (fTOT)


The total number of mandatory and spontaneous breaths per minute delivered
to the patient.

6.6.22

PAV-based Lung Compliance (CPAV)


For a PAV+ breath, CPAV is the change in pulmonary volume for an applied
change in patient airway pressure, measured under zero-flow conditions and

6-46

Operators Manual

Ventilator Settings

updated upon successful completion of each calculation. CPAV is displayed on


the waveform screen.
6.6.23

PAV-based Patient Resistance (RPAV)


For a PAV+ breath, RPAV is the change in pulmonary pressure for an applied
change in patient lung flow and updated upon successful completion of each
calculation. RPAV is displayed on the waveform screen.

6.6.24

PAV-based Lung Elastance (EPAV)


For a PAV+ breath, EPAV is the inverse of CPAV and is updated upon successful
completion of each calculation.

6.6.25

Spontaneous Rapid Shallow Breathing Index (f/ VT)


An indication of the patients ability to breathe spontaneously. High values
generally mean the patient is breathing rapidly, but with low tidal volumes.
Low values generally indicate the inverse.

6.6.26

Spontaneous Inspiratory Time Ratio (TI/TTOT)


In SPONT mode, the percentage of a spontaneous breath consumed by the
inspiratory phase. Updated at the successful completion of a spontaneous
breath.

6.6.27

Spontaneous Inspiratory Time (TI SPONT)


The duration of the inspiratory phase of a spontaneous breath and updated at
the end of each spontaneous breath. TI SPONT is only calculated when the
breathing mode allows spontaneous breaths and the breaths are patient-initiated.

6.6.28

PAV-based Total Airway Resistance (RTOT)


For a PAV+ breath, RTOT is the change in pulmonary pressure for an applied
change in total airway flow and updated upon the successful completion of

Operators Manual

6-47

Performance

each calculation. If the RPAV value appears in parentheses as described at the


beginning of this section, the RTOT value also appears in parentheses.
6.6.29

Static Compliance (CSTAT) and Static Resistance (RSTAT)


CSTAT is an estimate of the elasticity of the patients lungs, expressed in mL/
cmH2O. It is computed during a mandatory breath.
RSTAT is the total inspiratory resistance across the artificial airway and respiratory system, displayed at the start of the next inspiration after the Inspiratory
Pause maneuver. It is an estimate of how restrictive the patients airway is,
based on the pressure drop at a given flow, expressed in cmH2O/L/s. RSTAT is
computed during a VC mandatory breath with a square flow waveform.
CSTAT is calculated using this equation:

V pt
C STAT = ------------------------------ C ckt
P ckt PEEP

C STAT

Static compliance

P ckt

The pressure in the patient circuit measured at the end of the 100 ms interval
defining the pause-mechanics plateau

V pt

Total expiratory volume (patient and


breathing circuit)

PEEP

The pressure in the patient circuit measured at the end of expiration

C ckt

Compliance of the breathing circuit


during the pause maneuver (derived
from SST)

RSTAT is calculated using this equation after CSTAT is computed and assuming
a VC breath type with a SQUARE waveform:

R STAT

6-48

Static resistance

C STAT

Static compliance

Operators Manual

Ventilator Settings

R STAT

C ckt

V pt

C ckt
1 + -------------- P PEAK P PL
C STAT
= -------------------------------------------------------------------V pt

Compliance of the breathing circuit


during the pause maneuver (derived
from SST)

P PL

Flow into the patient during the last


100 ms of the waveform

P PEAK

Mean pressure in the patient circuit


over the 100 ms interval defining the
pause-mechanics plateau
Peak circuit pressure

During the pause, the most recently selected graphics are displayed and
frozen, to determine when inspiratory pressure stabilizes. CSTAT and RSTAT are
displayed at the start of the next inspiration following the inspiratory pause
and take this format:
CSTAT xxx
or
RSTAT yyy
Special formatting is applied if the software determines variables in the equations or the resulting CSTAT or RSTAT values are out of bounds:

6.6.30

Parentheses ( ) signify questionable CSTAT or RSTAT values, derived from questionable variables.

Flashing CSTAT or RSTAT values are out of bounds.

RSTAT ------ means resistance could not be computed, because the breath was not
of a mandatory, VC type with square flow waveform.

Dynamic Compliance (CDYN)


CDYN is a dynamic estimate of static compliance for each mandatory breath
delivered.

Operators Manual

6-49

Performance

6.6.31

Dynamic Resistance (RDYN)


RDYN is a dynamic estimate of static resistance for each mandatory breath
delivered.

6.6.32

C20/C
C20/C is the ratio of compliance of the last 20% of inspiration to the compliance of the entire inspiration.

6.6.33

End Expiratory Flow (EEF)


A measurement of the end expiratory flow for an applicable breath.

6.6.34

Peak Spontaneous Flow (PSF)


A measurement of the maximum inspiratory spontaneous flow for an applicable spontaneous breath

6.6.35

Displayed O2%
The percentage of oxygen in the gas delivered to the patient, measured at the
ventilators outlet, upstream of the inspiratory filter. It is intended to provide a
check against the set O2% for alarm determination, and not as a measurement
of oxygen delivered to the patient. O2% data can be displayed as long as the
O2 monitor is enabled. If the monitor is disabled, dashes (--) are displayed. If a
device alert occurs related to the O2 monitor, a blinking 0 is displayed.

6-50

Operators Manual

7 Preventive Maintenance

7.1

Overview
This chapter contains information on maintenance of the Puritan Bennett
980 Series Ventilator. It includes

7.2

How to perform routine preventive maintenance procedures, including frequency

How to clean, disinfect, or sterilize the ventilator and its main components

How to store the ventilator for extended periods

How to dispose of used parts

Ventilator Operational Time


The ventilator contains an hour meter that records the number of operational
hours since the ventilator was manufactured. An additional timer tracks the
number of hours since the last preventive maintenance activity was performed.
Both the GUI and the status display show the number of hours before the next
preventive maintenance is due.

7.3

Preventive Maintenance Intervals


WARNING:
To ensure proper ventilator operation, perform Preventive Maintenance
intervals as specified in the following tables. Reference Operator Preventive
Maintenance Frequency, p. 7-2 and Reference Service Preventive
Maintenance Frequency, p. 1-2 in the Operators Manual Addendum on
Preventive Maintenance.

7-1

Preventive Maintenance

Table7-1.Operator Preventive Maintenance Frequency


Part
Patient circuit: inspiratory and
expiratory limbs

Several times a day or as


required by the institution's
policy.

Maintenance

Check both limbs for


water accumulation.

Empty and clean.

Water traps, and drain bag

Check and empty as needed.

Oxygen sensor calibration

From the ventilator setup


screen, touch the More Settings
tab. To calibrate the oxygen
sensor, touch Calibrate in the
oxygen sensor area of the
screen.
Reference Oxygen sensor calibration testing, p. 4-34 for
information on testing the
oxygen sensor calibration.

Inlet air filter bowl

Replace bowl if it is
cracked.

If any sign of moisture is


visible, remove ventilator
from use and contact
service personnel.

Inspect and replace if


cracked, crazed, or damaged. Sterilize between
patients and circuit changes, or according to the
institution's policy. Sterilize before non-destructive
disposal, or dispose of
filter according to the institutions policy.

Run SST to check resistance of the inspiratory


limb.

Use care when changing


inspiratory filter to avoid
filter damage and minimize the potential for
introduction of particles.

Reusable inspiratory filter

7-2

Frequency

Before every use

After 15 days of continuous use in the inspiratory


limb (replace)

Yearly or after 50 autoclave cycles (replace)

Whenever excess resistance is suspected

Operators Manual

Service Personnel Preventive Maintenance

Table7-1.Operator Preventive Maintenance Frequency (Continued)


Part
Reusable expiratory filter

Condensate vial

Frequency

Before every use

After 15 days of continuous use in the exhalation


limb (replace)

Yearly or after 50 autoclave cycles (replace)

Whenever excess resistance is suspected

Several times a day or as


required by the institutions policy

Before every use

After 15 days of continuous use with the expiratory


filter (replace)

Maintenance

Inspect and replace if


cracked, crazed, or damaged. Sterilize between
patients and circuit changes, or according to the
institution's policy. Sterilize before non-destructive
disposal, or dispose of
filter according to the institutions policy.

Run SST to check resistance of the expiratory


limb and filter.

Use care when changing


expiratory filter to avoid
filter damage and minimize the potential for
introduction of particles.

Check and empty as


needed.

Inspect and replace if


cracked, crazed, or damaged.

Clean and sterilize


between patients and
circuit changes. Reference
Condensate Vial Cleaning,
p. 7-8 for instructions. Reference Component Sterilization, p. 7-24 for
sterilization requirements.

Sterilize before nondestructive disposal, or


according to the institutions policy.

Disposable inspiratory Filter

After 15 days of continuous use


(discard)

Discard according to the institutions protocol.

Disposable expiratory Filter

After 15 days of continuous use


(discard)

Discard according to the institutions protocol.

Operators Manual

7-3

Preventive Maintenance

Table7-1.Operator Preventive Maintenance Frequency (Continued)


Part

Frequency

Maintenance

Battery

When transferring battery to or


from another ventilator

Disinfect by wiping with a


damp cloth using one of the
solutions listed. Reference
Surface Cleaning Agents, p. 7-5
for approved cleaning agents.

Battery

Every three (3) years

Replace

Exhalation valve flow sensor


assembly

When contamination is suspected.

Reference Component Cleaning and Disinfection, p. 7-6 and


Reference Exhalation Valve
Flow Sensor Assembly Disinfection, p. 7-13.

NOTE: The exhalation valve


flow sensor assembly is removable and may be disinfected.
DO NOT STERILIZE the exhalation valve flow sensor assembly.
Exhalation valve flow sensor
assembly.

7.4

Every 25 disinfection cycles. A


disinfection cycle is defined as
one disinfection event as
described. Reference Exhalation Valve Flow Sensor Assembly Disinfection, p. 7-13.

Replace. Run flow sensor calibration, exhalation valve calibration, EST and SST after
replacing the exhalation valve
flow sensor assembly.

Caution:
Use specified cleaning, disinfecting, and sterilization agents and procedures
for the appropriate part as instructed.

Surface Cleaning of Exterior Surfaces


External surfaces of the GUI, BDU, and ventilator base may become soiled and
should be cleaned periodically.
To clean the GUI, BDU, and ventilator base surfaces
1.
Moisten a soft cloth with one of the surface cleaning agents listed or use SaniCloths (PDI, Inc.). Reference Surface Cleaning Agents, p. 7-5.

7-4

2.

Wipe the GUI, BDU, and ventilator base, removing any dirt or foreign substances.

3.

Dry all components thoroughly.

4.

If necessary, vacuum any cooling vents on the GUI and BDU with an electrostatic
discharge (ESD)-safe vacuum to remove any dust.

Operators Manual

Service Personnel Preventive Maintenance

Table7-2.Surface Cleaning Agents


Part
Ventilator exterior (including
touch screen and flex arm

Procedure

Comments/Cautions

Wipe clean with a cloth dampened with one of the cleaning


agents listed below or equivalent. Use a damp cloth and
water to rinse off chemical
residue as necessary.

Do not allow liquid or sprays to


penetrate the ventilator openings or cable connections.

Mild dish washing detergent


solution

Do not attempt to sterilize the


ventilator by exposure to ethylene oxide (ETO) gas.

Isopropyl alcohol (70% solution)

Do not use pressurized air to


clean or dry the ventilator,
including the GUI cooling vents.

Bleach (10% solution)

Do not submerge the ventilator


or pour cleaning solutions over
or into the ventilator

Window cleaning solution (isopropyl alcohol and ammonia)


Ammonia (15% solution)
Hydrogen peroxide (3% solution)
Formula 409 cleaner (Clorox
Company)
CaviCide surface disinfectant
(Metrex Research Corporation)
Control III germicide (Maril
Products, Inc.)
Virox (spray or pre-wipes) (Virox
Technologies Inc.)
Mr. Muscle Window & Glass
(SC Johnson
Sani Cloths (PDI, Inc.)
[Propan-2-ol, Isopropanol, Isopropyl Alcohol]1

Operators Manual

7-5

Preventive Maintenance

Table7-2.Surface Cleaning Agents (Continued)


Part
Ventilator cooling vents

Procedure
Vacuum the vents at the back
of the GUI and BDU to remove
dust.

Comments/Cautions
N/A

1. Chemicals stated are the generic equivalents of Mr. Muscle Window & Glass

7.5

Component Cleaning and Disinfection


WARNING:
To avoid microbial contamination and potential performance problems, do
not clean, disinfect, or reuse single-patient use (SPU) or disposable
components. Discard per local or institutional regulations.

Risks associated with reuse of single-patient use items include but are not
limited to microbial cross-contamination, leaks, loss of part integrity, and
increased pressure drop. When cleaning reusable components, do not use
hard brushes or implements that could damage surfaces.
Table7-3.Component Cleaning Agents and Disinfection Procedures
Part
Exhalation valve flow sensor
assembly

Cleaning Agent/Procedure

Comments/Cautions

Reference Exhalation Valve


Flow Sensor Assembly Disinfection, p. 7-13 for specific instructions.

Do not drop the exhalation


valve flow sensor assembly
or handle roughly during
disinfection or storage.

Cidex (2.5%) (ASP)

Sporox II (Sultan)

Cidex OPA (ASP)

.
Reusable patient circuit tubing

7-6

Disinfect per manufacturers


instructions-for-use.

Inspect for nicks and cuts,


and replace if damaged.

Run SST to check for leaks


when reinstalling the
circuit or when installing a
new circuit.

Operators Manual

Service Personnel Preventive Maintenance

Table7-3.Component Cleaning Agents and Disinfection Procedures (Continued)


Part

Cleaning Agent/Procedure

Comments/Cautions

Disposable patient circuit


tubing

Discard

Discard per the institutions


protocol.

Breathing circuit in-line water


traps

Disinfect per manufacturers


instructions-for-use.

Inspect water traps for


cracks and replace if damaged.

Run SST to check for leaks


when reinstalling the
circuit or when installing
new components.

Inspect components for


nicks and cuts, and replace
if damaged.

Run SST to check for leaks


when reinstalling the
circuit or when installing a
new components.

Breathing circuit components

Disinfect per manufacturers


instructions-for-use.

Disposable drain bag and


tubing (single unit)

Discard when filled to capacity


or when changing patient circuit.

N/A

Inlet air filter bowl

Wash the bowl with mild soap


solution, if needed.

Avoid exposing the inlet air


filter bowl to aromatic solvents, especially ketones.

Replace if cracks or crazing


are visible.

Battery

Wipe with a damp cloth using


one of the cleaning agents
listed. Reference Surface Cleaning Agents, p. 7-5.

Do not immerse the battery or


get the contacts wet.

Cooling fan filter

Clean every 250 hours or as


necessary.

Wash in mild soap solution,


rinse, and air dry.

Expiratory filter

Reference Expiratory Filter


Cleaning, p. 7-9 for cleaning
instructions.

Reference expiratory filter


instructions-for-use (IFU) for
sterilization instructions.

Operators Manual

Metrizyme (Metrex
Research Corporation)
ENZOL (ASP)

7-7

Preventive Maintenance

Table7-3.Component Cleaning Agents and Disinfection Procedures (Continued)


Part
Condensate vial

Cleaning Agent/Procedure
Reference Condensate Vial
Cleaning, p. 7-8 for cleaning
instructions.

Other accessories

Comments/Cautions
Clean condensate vial

Metrizyme (Metrex
Research Corporation)
ENZOL (ASP)

Follow manufacturers instructions-for-use.

N/A

To clean parts
1.
Wash parts in warm water using a mild soap solution.
2.

Thoroughly rinse parts in clean, warm water (tap water is acceptable) and wipe
dry.

3.

Clean or disinfect ventilator surfaces and component parts per the procedures
listed for each component. Reference Surface Cleaning Agents, p. 7-5. Reference
Component Cleaning Agents and Disinfection Procedures, p. 7-6 for acceptable
cleaning and disinfecting agents.

4.
5.

Visually inspect the components for cracks or other damage prior to use.
Dispose of damaged parts according to the institutions policy.

Whenever replacing or reinstalling a component, run SST before ventilating a


patient.
7.5.1

Condensate Vial Cleaning


Reference Condensate Vial, p. 7-9 when following the condensate vial cleaning instructions.

Note:
The o-ring (1) should remain in place during cleaning.
To clean the condensate vial
1.
Un-cap (2) the drain port from the condensate vial, but do not detach the retainer
(3).

7-8

Operators Manual

Service Personnel Preventive Maintenance

2.

Rinse the condensate vial in water to remove any accumulations prior to cleaning.

3.

Wipe the condensate vial assembly with the suggested cleaning solution to
remove any accumulated material. The vial assembly may be fully immersed in the
cleaning solution. Reference Component Cleaning Agents and Disinfection Procedures, p. 7-6 for the proper condensate vial cleaning solution.

4.

Rinse the vial assembly with clean water. The assembly may be fully immersed to
rinse.

5.

Wipe the condensate vial dry.


Figure7-1.Condensate Vial

7.5.2

O-ring

Drain cap

Retainer

Expiratory Filter Cleaning


Reference the photos in this section when cleaning the expiratory filter.

Operators Manual

7-9

liquid. To avoid retaining harmful residues, do not immerse housing or filter


medium in liquid or sterilize using ethylene oxide (ETO).
Preventive Maintenance

Caution:
Do not immerse the expiratory filter in liquid cleaning agent. Use care to
prevent the cleaning agent from making contact with the filter media inside
the filter housing.
Note:
Follow the institutions protocol for handling, storage, and disposal of potentially
biocontaminated waste.
To clean the expiratory filter
1.
Spray or wipe down surfaces of the exterior housing and housing bottom with
cleaning solution. Reference Component Cleaning Agents and Disinfection Procedures, p. 7-6, for appropriate cleaning solutions. Use care to prevent the solution
from contacting the filter media via the filter outlet, or entering the filter intake
opening or from patient port.

7-10

Operators Manual

Service Personnel Preventive Maintenance

Figure7-2. Filter Top and Bottom Views

Filter outlet (top)

External housing

Filter intake opening (bottom)

From patient port

Housing bottom

2.

Operators Manual

Wipe the interior surface of the from patient port with the solution to clean it.

7-11

Preventive Maintenance

Figure7-3.From Patient Port

3.

Allow to dry by orienting the filter in the upright position for the contact time
period recommended by the solutions manufacturer.

Figure7-4.Expiratory Filter Drying Orientation

7-12

Operators Manual

Service Personnel Preventive Maintenance

7.5.3

4.

Rinse all filter surfaces that were exposed to the cleaning solution with water,
using care to prevent liquid from entering the filter.

5.

Allow the filter housing to dry until all liquid has evaporated.

Exhalation Valve Flow Sensor Assembly Disinfection


Note:
Exhalation valve flow sensor assembly disinfection is not required on a routine basis
but it may be disinfected. Reference Component Cleaning Agents and Disinfection
Procedures, p. 7-6 for a list of appropriate disinfectants.
Note:
Follow the institutions infection control protocol for handling, storage, and disposal
of potentially bio-contaminated waste.
Caution:
To avoid damaging the hot film wire, do not insert fingers or objects into the
center port when disinfecting the exhalation valve flow sensor assembly.

The exhalation valve flow sensor assembly contains the exhalation valve body,
exhalation valve flow sensor, exhalation valve diaphragm, expiratory filter seal,
and pressure sensor filter. The expiratory flow sensor electronics consist of the
hot film wire and the thermistor. Since it is protected by the expiratory filter, it
does not require or need replacement or disinfection on a regular basis. It is,
however, removable and may be disinfected. Expected service life is 25 disinfection cycles.

Caution:
To avoid damage to the expiratory flow sensor element

Do not touch the hot film wire or thermistor in the center port

Do not vigorously agitate fluid through the center port while immersed.

Do not forcefully blow compressed air or any fluid into the center cavity.

Do not drop or handle roughly during disinfection or storage.

Operators Manual

7-13

Preventive Maintenance

7-14

WARNING:
Damaging the flow sensors hot film wire or thermistor in the center port can
cause the ventilators spirometry system to malfunction.

Operators Manual

Service Personnel Preventive Maintenance

Figure7-5.Exhalation Valve Flow Sensor Assembly

Top view

Bottom view

Figure7-6.Exhalation Valve Flow Sensor Components

Hot film wire and thermistor

Electrical contacts

Diaphragm sealing surface

Filter grommet

Removal
To remove the exhalation valve flow sensor assembly
1.
Lift up on the expiratory filter latch and open the expiratory filter door.
2.

Operators Manual

With thumb inserted into the plastic exhalation port and four (4) fingers under the
exhalation valve flow sensor assembly, pull it down until it snaps out. To avoid

7-15

Preventive Maintenance

damaging the flow sensor element, do not insert fingers into the center
port.
Figure7-7.Exhalation Valve Flow Sensor Assembly Removal

To remove disposable components of the exhalation valve flow sensor


assembly
1.
Remove and discard the exhalation valve diaphragm, the exhalation valve filter
seal, and the pressure sensor filter. Pinch the expiratory filter seal between two
fingers to remove it.
Figure7-8.Exhalation Valve Diaphragm Removal

7-16

Operators Manual

Service Personnel Preventive Maintenance

Figure7-9.Expiratory Filter Seal Removal

Figure7-10.Pressure Sensor Filter Removal

2.

Dispose of removed items according to the institutions protocol. Follow local governing ordinances regarding disposal of potentially bio-contaminated waste.

Disinfection

WARNING:
Do not steam autoclave the exhalation valve flow sensor assembly or sterilize
with ethylene oxide gas. Either process could cause the ventilators
spirometry system to malfunction when reinstalled in the ventilator.

Operators Manual

7-17

Preventive Maintenance

WARNING:
Use only the disinfectants described. Reference Component Cleaning Agents
and Disinfection Procedures, p. 7-6. Using disinfectants not recommended by
Covidien may damage the plastic enclosure or electronic sensor components,
resulting in malfunction of the ventilators spirometry system.
WARNING:
Follow disinfectant manufacturers recommendations for personal protection
(such as gloves, fume hood, etc.) to avoid potential injury.
1.
Pre-soak exhalation valve flow sensor assembly in the enzymatic solution.Reference Operator Preventive Maintenance Frequency, p. 7-2. The purpose for this
pre-soak is to break down any bio-film that may be present. Follow manufacturers instructions regarding duration of soak process.

7-18

Caution:
Do not use any type of brush to scrub the exhalation valve flow sensor
assembly, as damage to the flow sensing element could occur.
2.

Rinse in clean, de-ionized water.

3.

Prepare the chemical disinfectant according to the manufacturers instructions or


as noted in the institutions protocol. Reference Component Cleaning Agents and
Disinfection Procedures, p. 7-6 for the proper disinfecting agents.

4.

Immerse the in the disinfectant solution, oriented as shown, and rotate to remove
trapped air bubbles in the its cavities. Keep immersed for the minimum time
period by the manufacturer or as noted in the institutions protocol.

Operators Manual

Service Personnel Preventive Maintenance

Figure7-11.Immersion Method

5.

At the end of the disinfecting immersion period, remove and drain all disinfectant.
Ensure all cavities are completely drained.

Rinsing
1.

Rinse the exhalation valve flow sensor assembly using clean, de-ionized water in
the same manner used for the disinfection step.

2.

Drain and repeat rinsing three times with clean, de-ionized water.

WARNING:
Rinse according to manufacturers instructions. Avoid skin contact with
disinfecting agents to prevent possible injury.
3.

Operators Manual

After rinsing in de-ionized water, immerse in a clean isopropyl alcohol bath for
approximately 15 seconds. Slowly agitate and rotate to empty air pockets.

7-19

Preventive Maintenance

Drying
1.

Dry in a low temperature warm air cabinet designed for this purpose. Covidien
recommends a convective drying oven for this process, with temperature not
exceeding 60C (140F).

Caution:
Exercise care in placement and handling in a dryer to prevent damage to
the assemblys flow sensor element.

Inspection
Reference Exhalation Valve Flow Sensor Components, p. 7-15 while inspecting
the exhalation valve flow sensor assembly.

7-20

1.

Inspect the plastic body, diaphragm sealing surface, filter grommet and the seal
groove on the bottom side for any visible damage, degradation, or contamination.

2.

Inspect electrical contacts for contaminating film or material. Wipe clean with a
soft cloth if necessary.

3.

Inspect the hot film wire and thermistor in the center port for damage and for contamination.DO NOT ATTEMPT TO CLEAN EITHER OF THESE. If contamination
exists, rinse again with de-ionized water. If rinsing is unsuccessful or hot film wire
or thermistor is damaged, replace the exhalation valve flow sensor assembly.

Operators Manual

Service Personnel Preventive Maintenance

7.5.4

Exhalation Valve Flow Sensor Assembly Reassembly

The following illustration shows the reprocessing kit:


Figure7-12.Exhalation Valve Flow Sensor Assembly Reprocessing Kit

Diaphragm

Pressure sensor filter

1.

Expiratory filter seal

After drying the exhalation valve flow sensor assembly, remove the pressure
sensor filter from the reprocessing kit and install its large diameter into the filter
grommet with a twisting motion until flush with the plastic valve body, as shown.
The narrow end faces out.
Figure7-13.Installing the Pressure Sensor Filter

2.

Remove the expiratory filter seal from the kit and turn the assembly so its bottom
is facing up.

3.

Install the seal into the recess with the flat side facing outward, away from the
recess. Reference Installing the Expiratory Filter Seal, p. 7-22.

Operators Manual

7-21

Preventive Maintenance

Figure7-14.Installing the Expiratory Filter Seal

1
4.

7-22

Flat side of seal

Remove the diaphragm from the kit and install it. The outer seal bead rests in the
outer groove.

Operators Manual

Service Personnel Preventive Maintenance

Figure7-15.Installing the Diaphragm

1
5.

7.5.5

Diaphragm bead located in exhalation valve flow sensor assemblys groove

Carefully inspect component placement and the complete assembly.

Exhalation Valve Flow Sensor Assembly Replacement


1.

Replace the exhalation valve flow sensor assembly any time if cracked or damaged
in use, or if a malfunction occurs during SST or EST.

2.

Replace the assembly if damage is noted to the hot film wire and thermistor in the
center port.

3.

Perform required calibrations. Reference Operator Preventive Maintenance Frequency, p. 7-2.

To install the exhalation valve flow sensor assembly into the ventilator
1.
With the expiratory filter door open, insert the assembly directly under the exhalation valve and push straight up until it snaps into place. Reference Installing the
Exhalation Valve Flow Sensor Assembly, p. 7-24. To avoid damaging the hot film
wire, do not insert fingers into any opening.
2.

Install the expiratory filter by sliding it onto the tracks in the door, and orienting
the filters From Patient port through the hole in the door.

3.

Close expiratory filter door and lower expiratory filter latch.

Operators Manual

7-23

Preventive Maintenance

Figure7-16.Installing the Exhalation Valve Flow Sensor Assembly

4.

7.5.6

7.6

Run EST after disinfection.

Storage
1.

Pre-test the exhalation valve flow sensor assembly before storage by installing it
into the ventilator and running SST to test the integrity of the breathing system.
Reference To run SST, p. 3-48

2.

After performing SST, remove the assembly and place it into a protective bag or
similar covered container.

Component Sterilization
To sterilize parts
1.
Sterilize per the components instructions-for use or the steam sterilization procedure described. Reference Sterilization Parameters, p. 7-25 and Component Sterilization Procedures on page 7-25.

7-24

2.

After the components are sterilized, visually inspect them for cracks or other
damage.

3.

Dispose of damaged parts according to the institutions policy.

Operators Manual

Service Personnel Preventive Maintenance

Table7-4.Sterilization Parameters
Autoclave sterilization
Effective sterilization occurs by pre-vac sterilization of wrapped goods (132C for 4 minutes). Refer to
pre-vac system manufacturers program parameters or follow the steam sterilizer manufacturers
instructions.
1.

Disassemble the component.

2.

Clean the component.

3.

Wrap each component in a legally marketed sterilization wrap that has been cleared by FDA for
use in pre-vac sterilization using the parameters described above.

4.

Sterilize the wrapped parts*.

5.

Inspect the sterilized parts for damage, and discard if damaged.

6.

Reassemble the component.

7.

Install the component on the ventilator.

8.

Run SST.

*Follow pre-vac system manufacturers instructions for use (IFU).

Table7-5.Component Sterilization Procedures


Part

Procedure

Reusable expiratory and inspiratory filters

Steam autoclave per manufacturers instructions-for-use

Expiratory filter condensate vial

Operators Manual

Steam autoclave per manufacturers instructions-for-use

Comments/Cautions

Do not chemically disinfect


or expose to ETO gas

Check filter resistance


using ventilator SST or
other means before reuse.

Follow manufacturers recommendations for reuse.

Inspect the condensate vial


for cracks after processing.

Replace condensate vial if


damaged

7-25

Preventive Maintenance

Table7-5.Component Sterilization Procedures (Continued)


Part
Reusable drain bag tubing
(short piece of tubing attached
to drain bag) and clamp

Procedure
Clean and autoclave the reusable tubing; clean the clamp.
Reference Surface Cleaning
Agents, p. 7-5 for approved
cleaning agents.

Comments/Cautions
N/A

Whenever replacing or reinstalling a component, run SST before ventilating a


patient.
7.7

Service Personnel Preventive Maintenance


Covidien recommends only qualified service personnel perform preventive
maintenance activities. Complete details are described in the Puritan Bennett 980 Series Ventilator Service Manual.
At ventilator startup, and in Service mode, the GUI and status display indicate
when there are 500 hours or less before preventive maintenance is due.

7.8

Safety Checks
Covidien factory-trained service personnel should perform Extended Self Test
(EST) on the ventilator after servicing it at the intervals specified in the table
above. Reference the Puritan Bennett 980 Series Ventilator Service Manual
for details on performing EST.

7.9

Inspection and Calibration


Ventilator inspection and calibration should be performed by Covidien factorytrained service personnel at the intervals specified in the table above.

7.10

Documentation
Covidien factory-trained service personnel should manually enter the service
date, time, and nature of repair/preventive maintenance performed into the
log using a keyboard on the GUI.

7-26

Operators Manual

Service Personnel Preventive Maintenance

To manually document a service or preventive maintenance activity


1.
Enter Service Mode.

7.11

2.

Select the Logs tab.

3.

Select the Service Log tab.

4.

Select Add Entry, and using the buttons to the right of each line, complete the
entry.

5.

Touch Accept when complete.

Storage for Extended Periods


To store the ventilator
1.
Clean the unit thoroughly.
2.

Remove any batteries and accessories.

To return the ventilator to service


1.
Replace batteries.
2.

Recharge batteries prior to patient ventilation. If batteries are older than three (3)
years, use new batteries.

3.

Perform EST and SST prior to patient ventilation.

Operators Manual

7-27

Preventive Maintenance

Page Left Intentionally Blank

7-28

Operators Manual

8 Troubleshooting

8.1

Overview
This chapter contains information regarding ventilator logs on the Puritan Bennett 980 Series Ventilator.

8.2

WARNING:
To avoid a potential electrical shock, do not attempt to correct any electrical
problem with the ventilator while it is connected to AC power.

Problem Categories
For the Puritan Bennett 980 Series Ventilator Operators Manual, troubleshooting is limited to responding to ventilator alarms and reviewing various
ventilator logs. For detailed alarm information, including how to respond to
alarms, Reference Chapter 6 to address individual alarms that may occur
during ventilator use. Qualified service personnel who have attended the Covidien training class for Puritan Bennett 980 Series Ventilators should consult the
Puritan Bennett 980 Series Ventilator Service Manual for detailed repair
information and ventilator diagnostic codes.

8.3

How to Obtain Ventilator Service


To obtain service for the ventilator, call Covidien Customer Service at
1.800.635.5267 and follow the prompts.

8.4

Used Part Disposal


Follow local governing ordinances and recycling plans regarding disposal or
recycling of device components. Discard all damaged parts removed from the
ventilator during the maintenance procedures according to your institution's
protocol. Sterilize contaminated parts before non-destructive disposal.

8-1

Troubleshooting

8.5

Ventilator Logs
The ventilator uses various logs to store event information for later retrieval
when managing a patients treatment. Some of the logs are accessible during
ventilation and some logs are only available to Covidien personnel when the
ventilator is in Service mode. The Puritan Bennett 980 Series Ventilator
Service Manual gives more details regarding logs available to qualified service
personnel.
When New Patient is selected during ventilator setup, patient data, ventilator
settings, and alarm logs are cleared, but this information is available for Service
personnel review following New Patient selection when the ventilator is set up.

8-2

Alarms Log The alarm log records up to 1000 alarms that have occurred,
whether they have been reset or autoreset, the priority level, and their analysis
messages. The alarm log is accessible during normal ventilation and in Service
mode. A date- and time-stamped entry is made in the log whenever an alarm is
detected, escalated, reset or auto-reset. An entry is also made when an alarm
silence interval begins, ends, or is canceled. If one or more alarms have occurred
since the last time the alarm log was viewed, a triangular icon appears on the GUI
indicating there are unread items. The alarm log is stored in non-volatile memory
(NVRAM) and may be re-displayed after the ventilators power is cycled.If the ventilator enters BUV for any reason, this is also entered into the alarm log. The alarm
log is cleared by setting the ventilator up for a new patient.

Settings Log The settings log records changes to ventilator settings for retrospective analysis of ventilator-patient management. The time and date, old and
new settings. and alarm resets are recorded. A maximum of 500 settings changes
can be stored in the log. The settings log is cleared when the ventilator is set up
for a new patient. The settings log is accessible in normal ventilation mode and
Service mode.

Patient Data Log This log records every minute (up to 4320 patient data
entries) consisting of date and time of the entry, patient data name, and the
patient data value during ventilator operation. It is cleared when the ventilator is
set up for a new patient. Three tabs are contained in the patient data log:

Vital Patient data The log contains the same information that the clinician
has configured in the patient data banner at the top of the GUI. If the patient
data parameters in the banner are changed, these changes are reflected the
next time the patient data log is viewed.

Additional Patient Data 1 This log corresponds to the patient data parameters set on page 1 of the additional patient data banner. A total of 15 parameters are stored here, consisting of date and time of the entry (recorded every
minute), patient data name, and the patient data value during ventilator operation.

Operators Manual

Ventilator Logs

Additional Patient Data 2 This log corresponds to the patient data parameters set on page 2of the additional patient data banner. A total of ten (10)
parameters are stored here, consisting of date and time of the entry (recorded
every minute), patient data name, and the patient data value during ventilator
operation.

Diagnostic Log The Diagnostic Log is accessible during normal ventilation


and Service modes and contains tabs for the System Diagnostic Log (default), the
System Communication Log, and the EST/SST Diagnostic Log. The diagnostic log
contains tabs for the following:

System Diagnostic Log The System Diagnostic Log contains the date and
time when an event occurred, the type of event, the diagnostic code(s) associated with the event and any notes.Reference the Puritan Bennett 980
Series Ventilator Service Manual (10078090) for specific information contained in the System Diagnostic Log. The diagnostic log is not cleared when
the ventilator is set up for a new patient.

System Communication Log This log contains diagnostic information generated by the ventilators communication software. Reference the Puritan
Bennett 980 Series Ventilator Service Manual (10078090) for specific information contained in the System Communication Log.

EST/SST Diagnostic Log The EST/SST diagnostic log displays the time, date,
test/event, system code (reference the Puritan Bennett 980 Series Ventilator
Service Manual), type, and notes.

EST/SST status log The EST and SST status log displays the time, date, test,
test status (passed or failed).

General Event log The general event log contains ventilator-related information not found in any other logs. It includes changes in alarm volume, when the
ventilator entered and exited Stand-By, GUI key presses, respiratory mechanics
maneuvers, O2 calibration, patient connection, elevate O2, and warning notifications. The General event log can display up to 256 entries and is not cleared upon
new patient setup.

Service Log The service log is accessible during normal ventilation and Service
modes and contains the nature and type of the service, reference numbers specific
to the service event (for example, sensor and actuator ID numbers), manual and
automatic serial number input, and the time and date when the service event
occurred. It is not cleared upon new patient setup.

To view ventilator logs


1.
Touch the clipboard icon in the constant access icon area of the GUI. The log
screen appears with tabs for the various logs.
2.

Operators Manual

Touch the tab of the log desired.

8-3

Troubleshooting

3.

View the information for each parameter desired.


Figure8-1.Log Screen

Individual logs tabs

Pages contained in the log being viewed

Ventilator logs can be saved by entering Service mode, and downloading them
via the ethernet port. Reference the Puritan Bennett 980 Series Ventilator
Service Manual for instructions on downloading ventilator logs.
8.6

Diagnostic Codes
Refer to the diagnostic log for the codes generated during patient ventilation.
For a more information on the diagnostic codes, reference the Puritan Bennett 980 Series Ventilator Service Manual or contact Covidien Technical Support.

8-4

Operators Manual

9 Accessories

9.1

Overview
This chapter includes accessories that can be used with the Puritan Bennett
980 Series Ventilator. Reference Covidien Accessories and Options, p. 9-4 for
part numbers of any items available through Covidien.
The following commonly available accessories from the listed manufacturers
can be used with the ventilator system:

Filters DAR/Covidien, Puritan Bennett

Heated Humidification Systems Hudson RCI/Teleflex, Fisher & Paykel

Patient Circuits commonly available breathing circuits with standard ISO 15


mm/22 mm connection for neonatal, pediatric, and adult patients. Manufacturers
include Fisher & Paykel, DAR, and Hudson RCI/Teleflex.

Masks ResMed, Respironics, Fisher & Paykel

Patient Monitoring Systems Reference Connectivity to External Systems, p.


5-21 for information on which systems can be used with the ventilator

Nasal Interfaces Hudson RCI/Teleflex, Fisher & Paykel, Argyle

Compressed air filter and water trap Covidien

WARNING:
The Puritan Bennett 980 Series Ventilator contains phthalates. When used
as indicated, very limited exposure to trace amounts of phthalates may occur.
There is no clear clinical evidence that this degree of exposure increases
clinical risk. However, in order to minimize risk of phthalate exposure in
children and nursing or pregnant women, this product should only be used as
directed.

9-1

Accessories

9.2

General Accessory Information


The patient circuit support arm (flex arm) can be fastened to the ventilator
handle on either the right or left side. Flex arms used on the Puritan Bennett
840 Ventilator System can also be used on the Puritan Bennett 980 Ventilator System.

9-2

Operators Manual

General Accessory Information

Figure9-1.Ventilator with Accessories

Operators Manual

9-3

Accessories

Figure9-2.Additional Accessories

Reference Ventilator with Accessories, p. 9-3 and the figure above for the parts
listed in the following table.

Note:
Occasionally, part numbers change. If in doubt about a part number, contact your
local Covidien representative.
Table9-1.Covidien Accessories and Options
Item
number

9-4

Accessory or option description

Part number

Test lung

10005490

Drain Bag Tubing (package of 10)

4-048493-00

Drain Bag (package of 25)

4-048491-00

Drain Bag Tubing Clamp, reusable (package of 5)

4-048492-00

Pediatric-Adult Expiratory Filter1

10063033

Pediatric-Adult Expiratory Filtration System (carton of 12)

10043551

980 FRU, Exhalation valve flow sensor

10097468

Wall air water trap

10086051

Operators Manual

General Accessory Information

Table9-1.Covidien Accessories and Options (Continued)


Item
number
8

Operators Manual

Accessory or option description

Part number

Power cord, 10A, RA, ANZ

10087151

Power cord, 10A, RA, UK

10087159

Power cord, 10A, RA, EU

10087155

Power cord, 10A, RA, Japan

10087157

Power cord, 10A, RA, British

10087152

Power cord, 10A, RA, Switzerland

10087154

Power cord, 10A, RA, USA

10081056

Power cord, 10A, RA, Israel

10087156

Power cord, 10A, RA, Brazil

10087160

Power cord, 10A, RA, China

10087153

Air hose assembly; Norway, Sweden, Finland, Denmark,


Greece, France

4-074696-00

9-5

Accessories

Table9-1.Covidien Accessories and Options (Continued)


Item
number

Accessory or option description

Part number

Air hose assembly; Australia, New Zealand

4-074712-00

Air hose assembly; Canada

4-074709-00

Air hose assembly; Italy, Switzerland, Spain, Belarus, Kazakhstan

4-074706-00

Air hose assembly; Japan, Israel

10001777

Air hose assembly; Poland, Portugal, South Africa

4-074703-00

Air hose assembly; Switzerland

4-074707-00

Air hose assembly; United States, Latin America

4-006541-00

Air hose assembly; Germany, Luxembourg, Austria, Netherlands, Belgium, Croatia, Turkey, Russia, Slovenia, Serbia, Bulgaria, Romania

4-074714-00

Air hose assembly; United Kingdom, Ireland, Switzerland,


Hungary, Slovakia, Czech

4-074713-00

Oxygen hose assembly; Norway, Sweden, Finland, Denmark,


Greece, France

4-074697-00

Oxygen hose assembly; Canada

4-074710-00

Oxygen hose assembly, Italy, Switzerland, Spain, Belarus,


Kazakhstan

4-074705-00

Oxygen hose assembly; Japan, Israel

10001766

Oxygen hose assembly; Poland, Portugal, South Africa

4-074705-00

Oxygen hose assembly; Switzerland

4-074708-00

Oxygen hose assembly; United States, Latin America

4-001474-00

Oxygen hose assembly; Germany, Luxembourg, Austria,


Netherlands, Belgium, Croatia, Turkey, Russia, Slovenia,
Serbia, Bulgaria, Romania

4-074715-00

Oxygen hose assembly; United Kingdom, Ireland, Switzerland, Hungary, Slovakia, Czech

4-074698-00

For countries not identified, contact your local Covidien representative for the proper
air and oxygen hose part numbers.

9-6

10

Cylinder mount for compressed Air and O2 gas

10086050

11

Flex arm assembly

4-032006-00

12

Rechargeable Lithium Ion battery

10086042

Operators Manual

General Accessory Information

Table9-1.Covidien Accessories and Options (Continued)


Item
number

Accessory or option description

Part number

13

Humidifier bracket

10086049

14

Drain Bag Clip

10087137

Inspiratory bacterial filter, reusable (Re/Flex)

4-074600-00

Inspiratory bacterial filter, disposable, (carton of 12) (DAR)

351U5856

16

Condensate vial, reusable

10063031

17

Condensate vial drain cap

4-074613-00

Assy, Patient Circuit, Heated Wire, Adult (reusable)

G-061235-00

Assy, Patient Circuit, Heated Wire, Pediatric Patient circuit,


Pediatric, reusable

G-061237-00

Assy, Patient Circuit, Pediatric, Reusable

G-061223-00

Patient circuit, Adult, single use

6-003030-00

Ventilator breathing circuit, Neonatal disposable, DAR (not


available in USA)

3078447

Ventilator breathing circuit, Neonatal, disposable, Hudson


RCI/Teleflex

780-06

Ventilator breathing circuit, Neonatal, disposable, Fisher &


Paykel

RT235

19

O-ring seal, condensate vial, reusable

10085527

20

Neonatal expiratory filtration system, disposable, with condensate vial

4-076900-00

21

Proximal Flow monitoring sensor (disposable, 10/box)

10047078

Exhalation valve module reprocessing kit (6/carton)

10086048

15

18

Not shown

Hardware options
Not shown
--

Gold standard test circuit, 21 inch (for performing EST)

4-018506-00

Proximal Flow Monitoring Option

10084331

Software options

Operators Manual

Not shown

Leak Sync Software

10086418

Not shown

Leak Sync Software Upgrade

10096528

Not shown

IE Sync Software

10086354

Not shown

IE Sync Software Upgrade

10096529

9-7

Accessories

Table9-1.Covidien Accessories and Options (Continued)


Item
number

Accessory or option description

Part number

Not shown

NeoMode 2.0 Software

10086743

Not shown

NeoMode 2.0 Software Upgrade

10096526

Not shown

PAV+ Software

10086283

Not shown

PAV+ Software Upgrade

10096530

Not shown

BiLevel 2.0 Software

10086682

Not shown

BiLevel 2.0 Software Upgrade

10096527

1. Reusable filtration system does not include condensate vial. Reusable condensate vial must be ordered separately.

9-8

Operators Manual

10 Theory of Operations

10.1

Overview
This chapter provides specific details on breath delivery functions of the Puritan
Bennett 980 Series Ventilator. The chapter is organized as shown below.
Section Number

Title

Page

10.1

Overview

p. 10-1

10.2

Theoretical Principles

p. 10-3

10.3

Applicable Technology

p. 10-3

10.4

Inspiration Detection and initiation

p. 10-4

10.5

Exhalation Detection and Initiation

p. 10-9

10.6

Compliance and BTPS Compensation

p. 10-13

10.7

Mandatory Breath Delivery

p. 10-17

10.8

Spontaneous Breath Delivery

p. 10-22

10.9

A/C Mode

p. 10-31

10.10

SIMV Mode

p. 10-34

10.11

Spontaneous (SPONT) Mode

p. 10-39

10.12

Apnea Ventilation

p. 10-41

10.13

Detecting Occlusion and Disconnect

p. 10-45

10.14

Respiratory Mechanics

p. 10-49

10.15

Ventilator Settings

p. 10-57

10.16

Safety Net

p. 10-71

10.17

Power On Self Test (POST)

p. 10-76

10.18

Short Self Test (SST)

p. 10-77

10.19

Extended Self Test (EST)

p. 10-77

10-1

Theory of Operations

WARNING:
The ventilator offers a variety of breath delivery options. Throughout the
patient's treatment, the clinician should carefully select the ventilation mode
and settings to use for that patient based on clinical judgment, the condition
and needs of the patient, and the benefits, limitations and characteristics of
the breath delivery options. As the patient's condition changes over time,
periodically assess the chosen modes and settings to determine whether or
not those are best for the patient's current needs.

The gas supplies to which the ventilator are connected must be capable of
delivering 200 L/min flow with the supply pressure between 35 psig and 87
psig (241.8 kPa to 599.8 kPa). These supplies may be compressed air from an
external source (wall or bottled) air or oxygen.
Air and oxygen hoses connect directly to fittings at the rear of the breath delivery unit (BDU). The flow of each gas is metered by a Proportional Solenoid
(PSOL) valve to achieve the desired mix in the Mix Module. The flow through
each PSOL is monitored by separate flow sensors to ensure the accuracy of the
mix. The mixed gases then flow to the Inspiratory Module.
The blended gas in the Inspiratory Module is metered by the Breath Delivery
PSOL and monitored by the Breath Delivery Flow Sensor to ensure that the gas
is delivered to the patient according to the settings specified by the operator.
Delivered tidal volumes are corrected to standard respiratory conditions (BTPS)
to ensure consistent interpretation by the clinician. The Inspiratory Module also
incorporates the Safety Valve, which opens to vent excess pressure and allows
the patient to breathe room air (if able to do so) in the event of a serous malfunction.
Breathing gas exits the Inspiratory Module, passes through an internal bacteria
filter leaving the ventilator via the To Patient port. The ventilator breathing circuit, including the external bacteria filter and humidification means, carries the
breathing gas to the patient.
Exhaled gas leaves the expiratory limb of the breathing circuit through the
From Patient port. The gas is conducted through a condensate vial, the expiratory bacteria filter and the exhalation valve assembly (EVQ) which includes a
pressure sensing port, a flow sensor, and the exhaust port. The gas then flows
through the exhalation valve, which actively controls PEEP while minimizing
pressure overshoot and relieving excess pressures.
Pressure transducers in the inspiratory pneumatic system (PI) and exhalation
compartment (PE) monitor pressures for accurately controlling breath delivery.

10-2

Operators Manual

Short Self Test (SST)

10.2

Theoretical Principles
This theory of operations is described mainly from a clinical standpoint, discussing how the ventilator responds to various patient inputs, but also including a
general description of the ventilators components and how they work together to manage breath delivery.

10.3

Applicable Technology
The ventilators control is provided by Breath Delivery (BD) and Graphical User
Interface (GUI) Central Processing Units (CPUs). The BD CPU manages all
breath delivery functions and provides background checks on the subsystems
required for breath delivery. The GUI CPU controls the primary display, operator input devices, and the alarm system. The status display, a small, non-interactive LCD display located on the Breath Delivery Unit (BDU) is controlled by its
own processor. Reference Status Display, p. 2-28 for more information.
USB, Ethernet, and HDMI interfaces are provided on the ventilator. The USB
interface supports items such as transferring data to an external monitor via a
serial -over-USB protocol and saving screen captures to a memory storage
device or flash drive. Reference To configure Comm ports, p. 5-4 for information on serial-over-USB data transfer. The Ethernet interface is used by qualified service personnel for accessing ventilator logs and performing software
options installation, and the HDMI interface provides the ability to display the
GUI screen on an external video display device.
Pressure and flow sensors in the inspiratory and expiratory modules to manage
breath delivery processes. Sensor signals are used as feedback to the breath
delivery PSOL and exhalation valve controllers. Additional flow and pressure
sensors are used in the mix module to control the breathing gas composition.
In addition, gas temperature is measured for temperature compensation of
flow readings. Atmospheric pressure is measured in the inspiratory module
and used for BTPS compensation. The sensor signals are filtered using antialiasing filters and sampled with A/D converters. Additional low-pass filters
precondition the signals, the signals are then used for controls and display purposes.
Closed-loop control is used to maintain consistent pressure and flow waveforms in the face of changing patient/system conditions. This is accomplished
by using the output as a feedback signal that is compared to the operator-set
input. The difference between the two is used to drive the system toward the

Operators Manual

10-3

Theory of Operations

desired output. For example, pressure-control modes use airway pressure as


the feedback signal to control gas flow from the ventilator. Reference the
figure below. This diagram shows a schematic drawing of a general feedback
control system. The input is a reference value (e.g., operator preset inspiratory
pressure) that is compared to the actual output value (e.g., instantaneous
value of airway pressure). The difference between those two values is the error
signal. The error signal is passed to the controller (e.g., the software control
algorithm). The controller converts the error signal into a signal that can drive
the actuator (e.g., the hardware drivers and valves) to cause a change in the
manipulated variable (e.g., inspiratory flow).
Disturbances

Input

Error
signal

Controller
(software)

Actuator
(hardware)

Manipulated

Controlled
Plant

Variable

Variable

Feedback Signal

10.4

+
Note:
In the diagram
above, the plant is the patient and the connected breathing circuit.

Inspiration Detection and initiation


When ventilator inspiration occurs, it is called triggering. Breaths are delivered
to the patient based on ventilator settings the practitioner has entered and are
determined by pressure, flow, or time measurements, or operator action. The
ventilator uses the following methods to trigger an inspiration:

10-4

Pressure triggering (P-TRIG)

Flow triggering (V-TRIG)

Time-triggered

Operator-initiated

IE Sync triggered (if the IE Sync option is installed) (in SPONT mode, only)

Operators Manual

Short Self Test (SST)

Note:
Reference Appendix C in this manual for more information on IE Sync.

If the ventilator detects a drop in pressure at the circuit wye or when there is a
decrease in base flow measured at the exhalation valve, the patient is said to
trigger the breath. Mandatory breaths triggered by the patient are referred to
as PIM or patient-initiated mandatory breaths.
All spontaneous breaths are patient-initiated, and are also triggered by a
decrease in circuit pressure or measured base flow indicating the patient is initiating an inspiration.
Another term, autotriggering, is used to describe a condition where the ventilator triggers a breath in the absence of the patients breathing effort. Autotriggering can be caused by inappropriate ventilator sensitivity settings, water
in the patient circuit, or gas leaks in the patient circuit.
10.4.1

Pressure Triggering
If pressure triggering (P-TRIG) is selected, the ventilator transitions into inspiration when the pressure at the patient circuit wye drops below positive end
expiratory pressure (PEEP) minus the operator-set sensitivity level (PSENS). Reference the figure below. As the patient begins the inspiratory effort and
breathes gas from the circuit (event 5, the A-B interval in the figure, below),
pressure decreases below PEEP. When the pressure drops below PEEP minus
PSENS (event 6), the ventilator delivers a PIM breath. The pressure-decline time
interval between events A and B determines how aggressive the patients
inspiratory effort is. A short time interval signifies an aggressive breathing
effort. The A-B interval is also affected by PSENS. A smaller PSENS setting means
a shorter A-B time interval. (The minimum PSENS setting is limited by autotriggering, and the triggering criteria include filtering algorithms that minimize the
probability of autotriggering.)

Operators Manual

10-5

Theory of Operations

Figure10-1.Inspiration Using Pressure Sensitivity

10.4.2

Exhalation

Event B: Patient-triggered inspiration


begins

Inspiration

A-B interval

Event A: (patient inspires)

Operator-set pressure sensitivity

Flow Triggering
If flow triggering (V-TRIG) is selected the BDU provides a constant gas flow
through the ventilator breathing circuit (called base flow) during exhalation.
The base flow is 1.5 L/m greater than the value selected for flow sensitivity
(VSENS). Reference Inspiration Using Flow Sensitivity, p. 10-7 where the top
graphic represents expiratory flow and the bottom graphic represents inspiratory flow.]
The ventilators breath delivery flow sensor measures the base flow delivered
to the circuit and the exhalation valve flow sensor measures the flow entering
the exhalation valve. The ventilator monitors patient flow by measuring the difference between the inspiratory and exhaled flow measurements. If the
patient is not inspiring, any difference in measured flows is due to leaks in the
breathing system or flow sensor inaccuracy. The clinician can compensate for
leaks in the breathing system by increasing VSENSto a value equal to desired
VSENS + leak flow.

10-6

Operators Manual

Short Self Test (SST)

As the patient begins the inspiratory effort and inspires from the base flow, less
exhaled flow is measured, while the delivered flow remains constant. Reference the figure below (event A). As the patient continues to inspire, the difference between the delivery and exhalation valve flow sensor measurements
increases. The ventilator initiates an inspiration when the difference between
the two flow measurements is greater than or equal to the operator-set flow
sensitivity value. Reference Inspiration Using Flow Sensitivity, p. 10-7, (event B).
As with pressure triggering, the time delay between onset of the patients
effort and actual gas delivery depends on

How quickly the exhaled flow declines (that is, the aggressiveness of the inspiratory effort). The more aggressive the inspiratory effort, the shorter the interval,
and

the flow sensitivity value. The smaller the value, the shorter the delay

During flow triggering, a backup pressure sensitivity of 2 cmH2O is present to


detect a breath trigger in the event that the flow trigger fails.
Figure10-2.Inspiration Using Flow Sensitivity

Software-set base flow (L/min)

Operator-set flow sensitivity

Start of patient effort

1.5 L/min

Event A: flow is decreasing

Flow delivered to patient

Event B: Gas delivery begins

Operators Manual

10-7

Theory of Operations

10.4.3

Time Triggers
The ventilator measures the time interval for each breath and breath phase. If
the ventilator is in Assist/Control (A/C) mode (where the ventilator delivers
breaths based on the breath rate setting), a VIM or ventilator initiated mandatory breath is delivered after the appropriate time interval. The duration of the
breath in seconds (Tb) is 60/f.
Figure10-3.Breath Activity During Time-triggered Inspiration

10.4.4

Breath activity (VIM)

Breath activity (PIM)

Time period (Tb) = 60/f

Operator-initiated Triggers
If the operator presses the Manual inspiration key, an OIM (operator-initiated
mandatory) breath is delivered. The ventilator will not deliver an OIM under the
following conditions:

During an active inspiration, whether mandatory or spontaneous

During the restricted phase of exhalation

During circuit disconnect and Occlusion Status Cycling (OSC) conditions

Reference Manual Inspiration, p. 10-22 later in this chapter for information on


the restricted phase of exhalation.
10.4.5

IE Sync Trigger
IE Sync is a software option that allows the ventilator to trigger a breath based
on a model of the movement of the patients respiratory system. Reference
Appendix C in this manual for a detailed description of the IE Sync option.

10-8

Operators Manual

Short Self Test (SST)

10.5

Exhalation Detection and Initiation


When exhalation occurs, it is called cycling. Mandatory breaths can be volumecycled or time-cycled by the ventilator or pressure cycled by the patient. Spontaneous breaths can be flow-cycled or pressure-cycled by the patient or timecycled by the ventilator. A patient-cycled exhalation relies on measurements
such as inspiratory flow rate or airway pressure. The ventilator uses the three
(3) methods described below to detect exhalation:

10.5.1

Airway pressure method (spontaneous breaths)

Percent peak flow method (spontaneous breaths)

Time-cycling method (mandatory breaths)

IE Sync cycled (in SPONT mode only, if the IE Sync option is installed)

Airway Pressure Method


If expiratory sensitivity (ESENS) is set to a value too low for the patient-ventilator
combination, a forceful expiratory effort could cause circuit pressure (PPEAK) to
rise to its limit. The ventilator monitors circuit pressure throughout the inspiratory phase, and initiates an exhalation when the pressure equals the inspiratory
pressure (PI) target value + an incremental value. This transition to exhalation
occurs during spontaneous pressure-based ventilation and in volume support
(VS).

Note:
The allowable incremental value above the target pressure is 1.5 cmH2O once a
portion of inspiration time (Tn) has elapsed. Before Tn, the incremental value is higher
to allow for transient pressure overshoots. For the first 200 ms of inspiration, the
incremental pressure is 10% of the target pressure, up or 8 cmH2O, whichever is
greater. From 200 ms to Tn, the incremental pressure decreases in a linear fashion
from the initial value to 1.5 cmH2O.

Operators Manual

10-9

Theory of Operations

Figure10-4.Exhalation via the Airway Pressure Method

10.5.2

Pressure target

200 ms

Pressure target +incremental value (n)

Tn

Start breath

Percent Peak Flow Method


For spontaneous breath types including PS (pressure supported), TC (tube
compensated), and VS (volume supported, the ventilator captures the value of
the delivered peak inspiratory flow, then monitors the inspiratory flow decline
until the value of current flow to peak flow (expressed as a percentage) is less
than or equal to the set ESENS value. The ventilator then cycles from inspiration
into exhalation.
Reference Exhalation via the Percent Peak Flow Method, p. 10-11 for an
example of exhalation using this method.

10-10

Operators Manual

Short Self Test (SST)

Figure10-5.Exhalation via the Percent Peak Flow Method

10.5.3

Inspiratory flow (0 L/min)

Event B: Ventilator initiates exhalation

Inspiration

Inspiratory flow (L/min) without expiratory trigger

Trigger

VMAX x ESENS/100

Event A: delivered flow begins to


decrease (VMAX)

Note:
PAV+ uses a flow-based cycling method, also called ESENS, but it is expressed in L/min
rather than in % of VMAX.

Time-cycling Method
In pressure ventilation, the set inspiratory time (TI) defines the duration of the
inspiratory phase. In volume ventilation, TI depends on the tidal volume (VT)
setting, peak flow (VMAX), flow pattern, and plateau time (TPL). The ventilator
cycles into exhalation when the set TI (pressure ventilation) or computed TI
(volume ventilation) lapses.

10.5.4

IE Sync Cycling
IE Sync detects the end of inspiration when a real-time estimate of the timederivative of intra-pleural pressure,
threshold.

Operators Manual

(PSYNC), drops below the set ESENS

10-11

Theory of Operations

10.5.5

Backup Methods
There are four backup methods for preventing excessive duration or pressure
during inspiration

10.5.6

Time limit

High circuit pressure limit

High ventilator pressure limit

High inspired tidal volume limit

Time Limit
For adult and pediatric patients, the time limit method ends inspiration and
begins exhalation when the duration of a spontaneous inspiration is greater
than or equal to [1.99 s + 0.02 x PBW (kg)] s.

10.5.7

High Circuit Pressure Limit


During any type of inspiration, inspiration ends and exhalation begins when
the monitored airway pressure (PCIRC) is greater than or equal to the set high
circuit pressure limit.

10.5.8

High Ventilator Pressure Limit


The ventilator transitions from inspiration to exhalation if the high ventilator
pressure (2PVENT) limit of 110 cmH2O is reached.

10.5.9

High Inspired Tidal Volume Limit


The high inspired tidal volume limit terminates inspiration and commences
exhalation during VC+, VS, tube compensated (TC), or proportionally assisted
(PAV+) breaths if the delivered volume is greater than or equal to 2VTI.

10-12

Note:
The ventilator does not generate subatmospheric airway pressures during exhalation.

Operators Manual

Short Self Test (SST)

10.6

10.6.1

Compliance and BTPS Compensation


Compliance Compensation in Volume-based Breaths
Compliance compensation accounts for the gas volume not actually delivered
to the patient during inspiration. This gas is known as the compliance volume,
VC. VC is the gas lost to pressurizing the breathing circuit and includes the
volumes of the patient circuit, any accessories such as a humidifier and water
traps, and internal ventilator gas passages.
Figure10-6.Square Flow Pattern

Flow (y-axis)

Compliance volume (VC)

Actual VMAX

Set VT

Set VMAX

TI

Operators Manual

10-13

Theory of Operations

Figure10-7.Descending Ramp Flow Pattern

Flow (y-axis)

Set VT

Actual VMAX

TI

Set VMAX

Minimum VMAX

Compliance volume (VC)

In the ventilator, an iterative algorithm automatically computes the compliance


volume. There is a maximum tubing to- patient compliance ratio to reduce the
potential for over-inflation due to an erroneous patient compliance estimation.
The maximum ratio is determined by the selected patient circuit type and predicted body weight (PBW):

10-14

Operators Manual

Short Self Test (SST)

C pt ckt
Factor = --------------C pt

Factor

Compliance volume factor

C pt ckt

Compliance of the patient circuit

C pt

Compliance of the patient

The compliance volume is calculated as

V C = C pt ckt P wye P

VC

C pt ckt

Compliance volume

Compliance of the patient circuit

P wye

Pressure at the patient wye at the end of


the current inspiration

Pressure at the end of the current exhalation

Without automated compliance compensation, practitioners would have to


compute VC to estimate the loss of volume in the patient circuit, then increase
the VT setting by that amount. Increasing the tidal volume by a single increment to compensate for compliance volume provides only partial compensation, and requires extra effort and understanding by the practitioner.
Additionally, Pwye and P can change with time.

Operators Manual

10-15

Theory of Operations

An iterative algorithm in the ventilator automatically computes the compliance


volume and compensates for it. Compliance compensation does not change
inspiratory time (TI). It is achieved by increasing flow (increasing the amplitude
of the selected flow pattern). Keeping TI constant maintains the original I:E
ratio.
There is a maximum compliance volume to reduce the potential for overinflation due to an erroneous compliance volume calculation. The maximum compliance volume is determined by the selected patient circuit type and predicted
body weight (PBW), and is summarized by this equation:
Vcomp,max = Factor x Tidal volume
where:
Vcomp,max = maximum compliance volume
Factor = linear interpolation of the values in the following table for adult, pediatric, and neonatal circuit types. Factor is calculated as:
MIN (10, MAX (2.5, 1.0 + (2.0/0.3 x kg PBW)))
Table10-1.Compliance Volume Factors
Adult patient circuit type

10-16

Pediatric patient circuit type

PBW (kg)

Factor

PBW (kg)

Factor

10

10

15

4.6

11

3.5

30

3.4

12.5

2.9

60

2.75

15

2.7

150

2.5

30

2.5

Note:
Compliance compensation calculations are also in effect during exhalation to ensure
spirometry accuracy.

Operators Manual

Short Self Test (SST)

If the patients compliance decreases beyond the limits of compliance compensation, the ventilator relies on the 2PPEAK alarm setting to truncate the breath
and switch to exhalation.
10.6.2

BTPS Compensation in Volume-based Breaths


Volumes and flows are BTPS compensated, that is, they are reported by the
ventilator at existing barometric pressure, 37C (98.6F), and fully saturated
with water vapor.

10.7

Mandatory Breath Delivery


Three mandatory breath types are offered in the ventilator volume control
(VC) which bases breath delivery on the delivered inspiratory tidal volume,
pressure control (PC), which bases breath delivery on achieving and sustaining
a pressure target for a set period of time, and volume control plus (VC+) which
is a pressure-controlled breath based on a target tidal volume. VC+ can be
used in situations where a patients lungs become more compliant due to
treatment as it reduces the target pressure (lessening the forces on the alveoli)
to achieve the target tidal volume.
Mandatory breaths are delivered by the ventilator, are either assisted (if patient
initiated or PIM), or controlled (if ventilator initiated or VIM), or initiated by the
operator (OIM). In A/C mode, the breath period (Tb) is calculated using the
breath rate (f) according to the equation
T b = 60 f
If, during Tb, patient effort is detected, a PIM breath is initiated and a new
breath period starts. If no patient effort is detected before Tb lapses, the next
breath delivered is a VIM, and a new breath period starts.
Reference Ventilator Settings Range and Resolution, p. 11-8 for details on the
following VC+ settings:

Expiratory time (TE)

I:E ratio

Inspiratory time (TI)

Operators Manual

10-17

Theory of Operations

Rise time %

Target or tidal volume (VT)

VC and PC breath types require no initialization. A VC breath is based on


meeting a delivered volume target and a PC breath is based on meeting a specific pressure target. VC+ breaths, however, go through a startup routine.
10.7.1

Volume Control (VC)


Volume Control is the control scheme that controls the flow with for the
purpose of supplying a predetermined volume (set by the practitioner) to the
patient. There are two basic flow wave forms to administer this volume: the
square that guarantees a constant flow during the inspiration time, or the
descending ramp whose slope and initial value are determined to provide
the required volume target. Reference Ideal Waveform Using Square Flow Patternand Ideal Waveform Using Descending Ramp Flow Pattern. The inspiration
time is determined indirectly by the characteristics of the selected flow wave.
Figure10-8.Ideal Waveform Using Square Flow Pattern

10-18

Pressure (cmH2O)

Inspiration phase

Flow (L/min)

Expiration phase

Volume (mL)

Constant flow

Operators Manual

Short Self Test (SST)

Figure10-9.Ideal Waveform Using Descending Ramp Flow Pattern

10.7.2

Pressure (cmH2O)

Inspiration phase

Flow (L/min)

Expiration phase

Volume (mL)

Descending ramp

Pressure Control (PC)


Pressure Control is the control scheme by which the pressure is controlled at
the circuit wye to reach a constant level (set by the practitioner) during inspiration, and a PEEP level during exhalation. Reference Ideal Waveform Using
Pressure Control Ventilation on page 10-20. This level is maintained for a time
given by the set inspiration time, following followed by an exhalation regulated by the exhalation valve until the PEEP level is reached. As flow is not predetermined, the supplied volume varies depending on the patient's pulmonary
response.

Operators Manual

10-19

Theory of Operations

Figure10-10.Ideal Waveform Using Pressure Control Ventilation

10.7.3

Pressure (cmH2O)

PEEP

Flow (L/min)

Inspiration phase

Volume (mL)

Expiration phase

Target pressure

VC+
VC+ breaths require initialization and must go through a startup routine.
VC+ Startup
During VC+ startup, the ventilator delivers at least one breath (test breath) to
determine the pressure target needed to deliver the desired (set) volume.
During the time the ventilator is delivering the test breaths, the message VC+
startup is displayed in the GUIs prompt area.

10-20

Note:
To allow for optimal function of startup and operation of VC+ in the ventilator it is
important not to block the tubing while the patient is undergoing suctioning or other
treatment that requires disconnection from the ventilator. The ventilator has a
disconnect detection algorithm that suspends ventilation while the patient is
disconnected.

Operators Manual

Short Self Test (SST)

After VC+ Startup, the ventilator will make adjustments to the target pressure
in order to deliver the set volume (VT). In order to reach the desired volume
promptly, the maximum allowed pressure adjustments for an Adult or Pediatric
patient will be greatest during the first five breaths following Startup or a
change in VT or VT SUPP. The values of the maximum pressure adjustments for
each patient type are summarized below.

Table10-2.Maximum Pressure Adjustments


Conditions

Maximum change in target pressure


PBW 25 kg

15 PBW < 25 kg

PBW < 15 kg

Less than five breaths


after:
VC+ startup or
Change in VT

10.0 cmH2O

6.0 cmH2O

3.0 cmH2O

Five breaths or more


after VC+ startup

3.0 cmH2O

3.0 cmH2O

3.0 cmH2O

Reference Non-technical Alarm Summary, p. 6-18 for details on the following


VC+ alarms:

VOLUME NOT DELIVERED

HIGH INSPIRED TIDAL VOLUME (1VTI)

LOW CIRCUIT PRESSURE (3PPEAK)

COMPLIANCE LIMITED VT

During VC+, inspiratory target pressure cannot be lower than PEEP + 3 cmH2O
and cannot exceed PPEAK - 3 cmH2O.
10.7.4

Rise time %
If PC or VC+ is selected as the Mandatory type, adjust rise time % for optimum

flow delivery into the lungs. Patients with high impedance (low compliance,
high resistance) may benefit from a lower rise time% whereas patients with
low impedance may better tolerate a more aggressive rise time setting. The rise
time % setting specifies the speed at which the inspiratory pressure reaches

Operators Manual

10-21

Theory of Operations

95% of the target pressure. The rise time setting applies to PS (including a
setting of 0 cmH2O), PC, or VC+ breaths. To match the flow demand of an
actively breathing patient, observe simultaneous pressure-time and flow-time
curves, and adjust the rise time % to maintain a smooth rise of pressure to the
target value. A rise time % setting reaching the target value well before the
end of inspiration can cause the ventilator to supply excess flow to the patient.
Whether this oversupply is clinically beneficial must be evaluated for each
patient. Generally, the optimum rise time % for gently breathing patients is
less than or equal to the default (50%), while optimum rise time % for more
aggressively breathing patients can be 50% or higher.

10.7.5

WARNING:
Under certain clinical circumstances (such as stiff lungs, or a small patient with
a weak inspiratory drive), a rise time % setting above 50% could cause a
transient pressure overshoot and premature transition to exhalation, or
pressure oscillations during inspiration. Carefully evaluate the patients
condition before setting the rise time % above the default setting of 50%.

Manual Inspiration
When pressed, the Manual Inspiration key delivers one OIM breath to the
patient, using set breath delivery parameters. The ventilator will not allow a
manual inspiration during the restricted phase of exhalation or when the ventilator is in the process of delivering a breath (whether mandatory or spontaneous). All manual inspiration attempts are logged in the General Event log.
The restricted phase of exhalation is the time period during the exhalation
phase where an inspiration trigger is not allowed. The restricted phase of exhalation is defined as the first 200 ms of exhalation OR the time it takes for expiratory flow to drop to 50% of the peak expiratory flow, OR the time it takes
for the expiratory flow to drop to 0.5 L/min (whichever is longest). The
restricted phase of exhalation will end after five (5) s of exhalation have elapsed
regardless of the measured expiratory flow rate.

10.8

Spontaneous Breath Delivery


The modes allowing spontaneous breaths are SIMV, SPONT, and BiLevel.
The spontaneous breath type setting determines which type of pressure-assist
will be applied to the patients spontaneous breaths (PS, TC, VS, or PAV+).

10-22

Operators Manual

Short Self Test (SST)

After selecting the spontaneous breath type, choose the level of pressure
support (PSUPP) for PS, Support volume (VT SUPP) for VS or percent support for
TC and PAV+ (if the PAV+ option is installed) and specify the rise time % and
ESENS, where available. Changes to the spontaneous breath type setting phase
in at the start of the next inspiration.

Note:
In any delivered spontaneous breath, either INVASIVE or NIV, there is always a target
inspiratory pressure of at least 1.5 cmH2O applied.

During spontaneous breathing, the patient's respiratory control center rhythmically activates the inspiratory muscles. The support type setting allows selection of pressure-assist to supplement the patient's pressure-generating
capability.
Table10-3.Spontaneous Breath Delivery Characteristics
Characteristic

Implementation

Inspiratory detection

PSENS, VSENS, or IE Sync depending on the trigger


type selected.

Pressure or flow during inspiration


Spontaneous type = PS and PSUPP < 5 cmH2O

Pressure rises according to the selected rise time


% and PBW setting, with target pressure equal to
the effective pressure + PEEP:
PSUPP Effective pressure (cmH2O)
0
1.5
1
2.2
2
2.9
3
3.6
4
4.3

Pressure or flow during inspiration


Spontaneous type = PS and PSUPP 5 cmH2O

Pressure rises according to the selected rise time


% and PBW setting, and target pressure equals
PSUPP + PEEP.

Pressure or flow during inspiration


Spontaneous type = VS

Pressure rises according to the selected rise time


% and PBW setting, and target pressure equals
the pressure determined during the test breath or
pressure target determined from assessment of
delivered volume from the previous breath. For
more information on VS,Reference Volume
Support (VS), p. 10-25.

Operators Manual

10-23

Theory of Operations

Table10-3.Spontaneous Breath Delivery Characteristics (Continued)


Characteristic

10-24

Implementation

Tube Compensation (TC)

Tube Compensation provides programmable,


inspiratory pressure assistance during otherwise
unsupported spontaneous breaths. This assists
the patient in overcoming the flow resistance of
the artificial airway. Pressure is programmed to
help the patient overcome part or all of the resistance of the artificial airway. The ventilator continuously calculates the pressure differential
based on tube type and tube ID and adjusts the
compensation pressure accordingly. For more
information regarding TC, Reference Tube Compensation, p. 10-27.

Inspiratory flow profile

The inspiratory flow profile is determined by


patient demand and the rise time % setting. As
the rise time % setting is increased from
minimum to maximum, the time to achieve the
pressure target decreases. The maximum available flow is up to 30 L/min for neonatal circuit
types, 80 L/min for pediatric circuit types, and up
to 200 L/min for adult circuit types without Leak
Sync.

Exhalation valve during inspiration

Adjusts to minimize pressure overshoot and


maintain the target pressure.

Inspiratory valve during inspiration

Adjust to maintain target pressure.


Because the exhalation valve acts as a relief valve
venting any excess flow, inspiratory flow can be
delivered aggressively and allows reduced work
of breathing.

Expiratory detection

The end-inspiratory flow or airway pressure


method or IE Sync cycling (if IE Sync is installed),
whichever detects exhalation first. Time backup
and the 1PPEAK alarm are also available as backup
strategies.

Pressure or flow during exhalation

Pressure is controlled to PEEP.


For pressure triggering: set to deliver a bias flow
of 1 L/min near the end of expiratory flow.
For flow triggering: set to deliver base flow.

Inspiratory valves during exhalation

For pressure triggering: set to deliver a bias flow


of 1 L/min near the end of expiratory flow.
For flow triggering: set to deliver base flow near
the end of expiratory flow.

Exhalation valve during exhalation

Adjusts to maintain the operator-selected value


for PEEP.

Operators Manual

Short Self Test (SST)

10.8.1

Pressure Support (PS)


Pressure Support is a type of spontaneous breath, similar to PC, by which the
pressure is controlled to reach a constant value, preset by the practitioner,
once an inspiratory effort is detected. This target value is held until the detection of end of inspiration. Subsequently, the exhalation valve control initiates
the exhalation, driving the pressure to the PEEP level.

10.8.2

Volume Support (VS)


Volume support is a pressure-supported spontaneous breath type available
when SPONT is selected as the mode. The target support volume (VT SUPP) is
the target volume for pressure supported breaths.
Reference Ventilator Settings Range and Resolution, p. 11-8 for details regarding the following VS settings:

Expiratory sensitivity (ESENS)

Rise time %

Target support volume (VT SUPP)

Technical Description
Volume Support (VS) breaths are patient-triggered, pressure-supported spontaneous breaths. The VS algorithm varies the inspiratory pressure of each
breath to deliver the operator-set target tidal volume (VT SUPP). If the delivered
volume for a breath is above or below the set target volume, VS adjusts the
target pressure for the next breath up or down, as necessary, to deliver more
or less volume. As the patient's condition improves allowing more patient
control over spontaneous ventilation, the VS algorithm decreases the amount
of inspiratory pressure necessary to deliver the target volume. Conversely, VS
increases inspiratory pressure if the patient's respiratory drive becomes compromised.
In the absence of leaks or changes in patient resistance or compliance, Volume
Support achieves and maintains a steady, breath-to-breath tidal volume within
five (5) breaths of VS initiation or startup.
During VS, the inspiratory pressure target cannot be lower than PEEP + 1.5
cmH2O, and cannot exceed PPEAK - 3 cmH2O.

Operators Manual

10-25

Theory of Operations

VS Startup
During startup, the ventilator delivers a breath (test breath) to determine the
pressure target needed to deliver the desired (set) volume. During the time the
ventilator is delivering the test breath, the message VS startup is displayed
in the GUIs prompt area.
Test breaths are defined as:

A PS spontaneous breath with PSUPP = the lesser of 15 cmH2O and (PPEAK-PEEP-3)


cmH2O
Note:
To allow for optimal function of startup and operation of VS in the ventilator it is
important not to block the tubing while the patient is undergoing suctioning or
other treatment that requires disconnection from the ventilator. The ventilator has
a disconnect detection algorithm that suspends ventilation while the patient is
disconnected.

After VS Startup, the ventilator makes adjustments to the target pressure in


order to deliver the set volume (VT SUPP). In order to reach the desired volume
promptly, the maximum allowed pressure adjustments for an Adult or Pediatric
patient will be greatest during the first five breaths following Startup or a
change in VT SUPP. The values of the maximum pressure adjustments for each
patient type are summarized in the table below.
Table10-4.Maximum Pressure Adjustments
Conditions

Maximum change in target pressure


PBW 25 kg

15 kg PBW < 25 kg

PBW < 15 kg

Less than five breaths


after:
VS startup or Change
in VT SUPP

10.0 cmH2O

6.0 cmH2O

3.0 cmH2O

Five breaths or more


after VS startup

3.0 cmH2O

3.0 cmH2O

3.0 cmH2O

Reference Non-technical Alarm Summary, p. 6-18 for details on the following


VS alarms:

10-26

VOLUME NOT DELIVERED

COMPLIANCE LIMITED VT

HIGH INSPIRED TIDAL VOLUME (1VTI)

Operators Manual

Short Self Test (SST)

Monitored Patient Data


Reference Patient Data Range and Resolution, p. 11-19 for details on the Spontaneous inspired tidal volume patient data parameter available during VS
breaths.
10.8.3

Tube Compensation
Tube Compensation (TC) is a pressure-supported spontaneous breath type
available in SIMV, SPONT and BiLevel modes. When TC is enabled, the patients
respiratory muscles are not required to work as hard to draw gases into the
lungs as they would in the absence of the pressure assistance provided by the
TC feature. This is particularly important for patients whose respiratory systems
are already functioning poorly, and would have to exert even greater muscular
effort to overcome the increased resistance to flow through the artificial
airway.
Tube Compensation provides programmable, inspiratory pressure assistance
during otherwise unsupported spontaneous breaths. This assists the patient in
overcoming the flow resistance of the artificial airway. Pressure is programmed
to vary in accordance with the resistance to flow of the artificial airway. The
ventilator continuously calculates the pressure differential and adjusts the
compensation pressure accordingly.
Tube Compensation also includes safety protection, safety checks, and logic
checks which prevent the operator from entering certain incompatible settings, such as a large airway size paired with a small predicted body weight.
If the type of humidifier has been changed after running SST with TC, the
volume can be adjusted at the same time to avoid a reduction in compensation
compliance accuracy.
Technical Description
Tube Compensation is a spontaneous mode enhancement which assists
patients spontaneous breaths not already supported by specific pressurebased breath types (such as PS, VS, and PAV+) by delivering positive pressure
proportional to the flow-based, resistive pressure developed across the artificial airway. TC causes the sensation of breathing through an artificial airway
to diminish because the TC algorithm instructs the ventilator to develop just
the correct amount of forward pressure to offset (cancel) the back pressure
developed across the artificial airway during the inspiratory phase. The degree

Operators Manual

10-27

Theory of Operations

of cancellation can be set by the clinician and is adjustable between 10% an


100% in increments of 5%.
Tube Compensation can support all unsupported spontaneous breaths for
patients with predicted body weights 7.0 kg (15.4 lb), and for endotracheal/
tracheostomy tubes with an inside diameter (ID) of 4.5 mm. TC can be used
within SPONT, BiLevel (if this option is installed) or SIMV, all of which permit
unsupported spontaneous breaths. With BiLevel selected, TC supports spontaneous breaths at both pressure levels.
Tube Compensation checks the flow rate every 5 ms, using an internal lookup
table which contains the flow-to-pressure relationship of the selected artificial
airway, and is used to calculate the amount of pressure needed to overcome
all or part of the resistance of the artificial airway. Based on the TC setting and
the instantaneous flow measurement, the ventilators PSOL valves are continually adjusted, adjusting the circuit pressure to match the changing tube-pressure compensation requirements.
Tube Compensation Alarms
Reference Non-technical Alarm Summary, p. 6-18 for details of the 1PCOMP,
1PVENT, and 1VTI alarms associated with TC.
Monitored Patient Data
Reference Patient Data Range and Resolution, p. 11-19 for details of the
inspired tidal volume (VTI) monitored patient data parameter a associated with
TC.
Tube Inside Diameter (ID)
The ventilator uses soft bound and hard bound values for estimated tube
inside diameter (ID) based on PBW. Soft bounds are ventilator settings that
have reached their recommended high or low limits. When adjusting the tube
size, if the inside diameter does not align with a valid predicted body weight,
a Continue button appears. Setting the ventilator beyond these soft bounds
requires the operator to acknowledge the prompt by touching the Continue
button before continuing to adjust the tube size.The limit beyond which the
tube ID cannot be adjusted is called a hard bound, and the ventilator emits an
invalid entry tone when a hard bound is reached.

10-28

Operators Manual

Short Self Test (SST)

WARNING:
Greater than expected ventilatory support, leading to unknown harm, can
result if the specified tube type or tube ID is smaller than the actual tube type
or tube ID.

Ventilator Settings/Guidelines
The estimation of settings to use with TC is aided by an understanding of: the
ventilator settings, the data used for determination of the compensation
values, and the specified performance or accuracy of the TC function.
The setting for 2PPEAK must take the estimated tube compensation into consideration. The target pressure (compensation) at the patient wye is derived
from the knowledge of the approximate airway resistance of the ET or tracheostomy tube being used. The compensation pressure in cmH2O for available
tube sizes and gas flows is shown. Reference ET Tube Target Pressure vs. Flow,
p. 10-30 and Reference Tracheostomy Tube Target Pressure vs. Flow, p. 10-31.
The estimated compensation must be added to the value of PEEP for calculation and setting of 2PPEAK.
Specified Performance
Performance using TC is specified to be (0.5 + 10% of actual) joules/liter
(residual work during inspiration at the 100% support (% Supp) level). Work
is computed over the entire inspiratory interval.

k P E END P TR V dt
W = -----------------------------------------------------------------V dt

Work [J/L]

PTR

Tracheal pressure

PE END

End expiratory pressure

Conversion constant (0.098) [J/cmH2O x L)

Operators Manual

10-29

Theory of Operations

The following figures indicate pressures at steady-state flows for ET tubes and
tracheostomy tubes, respectively, at 100% support at the wye for sizes
between 4.5 mm and 10 mm.
Figure10-11.ET Tube Target Pressure vs. Flow

10-30

Pressure (cmH2O)

Flow (L/min)

Operators Manual

Short Self Test (SST)

Figure10-12.Tracheostomy Tube Target Pressure vs. Flow

10.8.4

Pressure (cmH2O)

Flow (L/min)

Proportional Assist Ventilation (PAV+)


PAV+ is another type of spontaneous breath, which is available only if the
PAV+ option is installed. For detailed description of the operating theory, reference Appendix D in this manual.

10.9

A/C Mode
When the ventilator is in assist-control (A/C) mode, only mandatory breaths
are delivered. These mandatory breaths can be PC, VC, or VC+ breaths. Reference Mandatory Breath Delivery, p. 10-17 for a more detailed explanation of
VC+ breaths. As for any mandatory breath, the triggering methods can be
P-TRIG, V-TRIG, time-triggered, or operator initiated. If the ventilator senses the
patient initiating the breath, a PIM or assist breath is delivered. Otherwise,
VIM breaths (control breaths) are delivered based on the set respiratory rate.
The length of the breath period is defined as

Operators Manual

10-31

Theory of Operations

where:

Tb = 60 f

Tb = breath period (s)


f = set respiratory rate (breaths per minute)
The inspiratory phase length is determined by the current breath delivery settings. At the end of the inspiratory phase, the ventilator enters the expiratory
phase as determined by the following equation:

T E = Tb T I

where:
TE = length of the expiratory phase (s)
TI = length of inspiratory phase(s) including plateau time, TPL
The figure shown below illustrates A/C breath delivery when there is no patient
inspiratory effort detected (all inspirations are VIMs).
Figure10-13.No Patient Inspiratory Effort Detected

10-32

VIM

Tb

Operators Manual

Short Self Test (SST)

The figure below shows A/C breath delivery when patient inspiratory effort is
detected. The ventilator allows PIM breaths to be delivered at a rate greater
than or equal to the set respiratory rate.
Figure10-14.Patient Inspiratory Effort Detected

PIM

Tb set

The figure shown below illustrates A/C breath delivery when there are both
PIM and VIM breaths. delivered.
Figure10-15.Combined VIM and PIM Breaths

VIM

PIM

Tb set

If changes to the respiratory rate are made, they are phased in during exhalation only. The new breath period depends on the new respiratory rate, is based
on the start of the current breath, and follows these rules

The current breaths inspiratory time is not changed.

A new inspiration is not delivered until at least 200 ms of exhalation have elapsed.

The maximum time t until the first VIM for the new respiratory rate is delivered is
3.5 times the current inspiratory time or the length of the new breath period
(whichever is longer), but t is no longer than the old breath period.

If the patient generates a PIM after the ventilator recognizes the rate change and
before time t, the new rate begins with the PIM.

Operators Manual

10-33

Theory of Operations

10.9.1

Changing to A/C Mode


Switching to A/C mode from any other mode causes the ventilator to phase in
a VIM and set the start time for the beginning of the next A/C breath period.
Following this VIM, and before the next A/C period begins, the ventilator
responds to the patients inspiratory efforts by delivering mandatory breaths.
The first A/C breath (VIM breath) is phased in while following these rules

The breath is not delivered during an inspiration.

The breath is not delivered during the restricted phase of exhalation.

The ventilator ensures the apnea interval elapses at least five (5) s after the beginning of exhalation.

Any other specially scheduled event (for example, a respiratory mechanics maneuver or any pause maneuver) is canceled and rescheduled at the next interval.

When the first VIM of the new A/C mode is delivered depends on the mode
and breath type active when the mode change is requested.
10.10

SIMV Mode
Synchronous Intermittent Mandatory Ventilation (SIMV) mode is a mixed ventilation mode allowing both mandatory and spontaneous breaths using pressure- or flow-triggering. The mandatory breaths can be PC, VC, or VC+, and
the spontaneous breaths are pressure-assisted with either PS or TC. SIMV guarantees one mandatory breath per SIMV breath period, which is either a PIM or
VIM. OIM breaths are allowed in SIMV and are delivered at the setting selected
for Mandatory Type. Reference the figure below which shows the two parts of
the SIMV breath period.
Figure10-16.Mandatory and Spontaneous Intervals

1 Tb = SIMV breath period (includes Tm and Ts 3 Ts = Spontaneous interval (VIM delivered if no


PIM delivered during Tm
2 Tm = Mandatory interval (reserved for a PIM
breath)

10-34

Operators Manual

Short Self Test (SST)

The first part of the period is the mandatory interval (Tm) which is reserved for
a PIM. If a PIM is delivered, the Tm interval ends and the ventilator switches to
the second part of the period, the spontaneous interval (Ts), which is reserved
for spontaneous breathing for the remainder of the breath period. At the end
of an SIMV breath period, the cycle repeats. If a PIM is not delivered during the
mandatory interval, the ventilator delivers a VIM at the end of the mandatory
interval, then switches to the spontaneous interval. The following figure shows
an SIMV breath period where a PIM is delivered within the mandatory interval.
Any subsequent trigger efforts during Ts yield spontaneous breaths. As shown,
Tm transitions to Ts when a PIM is delivered.
Figure10-17.PIM Delivered Within Mandatory Interval

PIM

Ts (subsequent trigger efforts


during Ts yield spontaneous
breaths)

Tm (Tm transitions to Ts when a


PIM is delivered)

Tb

The following figure shows an SIMV breath period where a PIM is not delivered
within the mandatory interval.
Figure10-18.PIM Not Delivered Within Mandatory Interval

Operators Manual

VIM

Ts

10-35

Theory of Operations

Tm (VIM delivered at end of Tm


if no PIM delivered during Tm

Tb

In SIMV, mandatory breaths are identical to those in A/C mode if the ventilators respiratory rate setting is greater than the patients natural respiratory
rate. Spontaneous breaths are identical to those in SPONT mode if the ventilator setting for respiratory rate is significantly below the patients natural respiratory rate. Patient triggering must meet the requirements for pressure and
flow sensitivity.
The procedure for setting the respiratory rate in SIMV is the same as in A/C
mode. Once the respiratory rate (f) is set, the SIMV interval period Tb in
seconds is
Tb = 60 f
During the mandatory interval, if the patient triggers a breath according to the
current setting for pressure or flow sensitivity, the ventilator delivers a PIM.
Once a mandatory breath is triggered, Tm ends, Ts begins, and any further
trigger efforts yield spontaneous breaths. During the spontaneous interval, the
patient can take as many spontaneous breaths as allowed. If no PIM or OIM is
delivered by the end of the mandatory interval, the ventilator delivers a VIM
and transitions to the spontaneous interval at the beginning of the VIM.
The SIMV breathing algorithm delivers one mandatory breath each period
interval, regardless of the patients ability to breathe spontaneously. Once a
PIM or VIM is delivered, all successful patient efforts yield spontaneous breaths
until the cycle interval ends. The ventilator delivers one mandatory breath
during the mandatory interval, regardless of the number of successful patient
efforts detected during the spontaneous interval. (An OIM delivered during the
mandatory interval satisfies the mandatory breath requirement, and causes Tm
to transition to Ts.)
The maximum mandatory interval for any valid respiratory rate setting in SIMV
is defined as the lesser of

0.6 x the SIMV interval period (Tb), or

ten s.

There is no minimum value for Tm.


In SIMV, the interval from mandatory breath to mandatory breath can be as
long as 1.6 x the SIMV period interval (but no longer than the period interval

10-36

Operators Manual

Short Self Test (SST)

+ ten (10) s. At high respiratory rates and too-large tidal volumes, breath
stacking (the delivery of a second inspiration before the first exhalation is
complete) is likely. In volume ventilation, breath stacking during inspiration
and early exhalation leads to hyperinflation and increased airway and lung
pressures, which can be detected by a high pressure limit alarm. In pressure
control ventilation (with inspiratory pressure remaining constant), breath
stacking leads to reduced tidal volumes, which can be detected by the low tidal
volume and minute ventilation alarms.
In SIMV mode it is possible for the respiratory rate to drop temporarily below
the f setting (unlike A/C mode, in which fTOT is always greater than or equal to
the f setting). If the patient triggers a breath at the beginning of a breath
period, then does not trigger another breath until the maximum mandatory
interval for the following breath has elapsed, a monitored respiratory rate less
than the respiratory rate setting can result.
If a spontaneous breath occurs toward the end of the spontaneous interval,
inspiration or exhalation can still be in progress when the SIMV interval ends.
No VIM, PIM, or OIM is allowed during the restricted phase of exhalation. In
the extreme, one or more expected mandatory breaths could be omitted.
When the expiratory phase of the spontaneous breath ends, the ventilator
reverts to its normal criteria for delivering mandatory breaths.
If an OIM is detected during the mandatory interval, the ventilator delivers the
currently specified mandatory breath then closes Tm and transitions to Ts. If an
OIM is detected during the spontaneous interval, the ventilator delivers the
currently specified mandatory breath, but the SIMV cycle timing does not
restart if OIM breaths are delivered during Ts.
10.10.1

Changing to SIMV Mode


Switching the ventilator to SIMV from any other mode, causes the ventilator
to phase in a VIM and set the start time for the next SIMV period. Following
this VIM, but before the next SIMV period begins, the ventilator responds to
successful patient inspiratory efforts by delivering spontaneous breaths. The
first SIMV VIM breath is phased in according to the following rules:

The VIM breath is not delivered during an inspiration or during the restricted
phase of exhalation.

If the current mode is A/C, the first SIMV VIM is delivered after the restricted phase
of exhalation plus the shortest of the following intervals, referenced to the beginning of the last or current inspiration: 3.5 TI, current TA, or the length of the
current breath period.

Operators Manual

10-37

Theory of Operations

If the current mode is SPONT, and the current or last breath type was spontaneous
or OIM, the first SIMV VIM is delivered after the restricted phase of exhalation plus
the shortest of the following intervals, referenced to the beginning of the last or
current inspiration: 3.5 x TI, or current TA.

If the current mode is BiLevel in the PH state and the current breath is mandatory,
the PEEP level will be reduced to PL once the exhalation phase is detected

The time t until the first VIM of the new A/C mode is the lesser of:

PEEP transition time + 2.5 x duration of the active gas delivery phase, or

the length of the apnea interval (TA), or

the length of the current breath cycle

If the current mode is BiLevel in the PH state and the current breath is spontaneous:

the PEEP level will be reduced once the exhalation phase is detected

The time t until the first VIM of the new A/C mode is the lesser of:

10-38

PEEP transition time + 2.5 x duration of the spontaneous inspiration, or

the start time of the spontaneous breath + the length of the apnea interval
(TA).

If the current mode is BiLevel in the PL state and the current breath is mandatory,
the time t until the first VIM of the new A/C mode is the lesser of:

PEEP transition time + 2.5 x duration of the active gas delivery phase, or

the length of the apnea interval (TA), or

the length of the current breath cycle

If the current mode is BiLevel in the PL state and the current breath is spontaneous
and the spontaneous start time has occurred during PL, the time t until the first
VIM of the new A/C mode is the lesser of:

3.5 x duration of the spontaneous inspiration, or

the length of the apnea interval (TA) or

the length of the current breath cycle

Operators Manual

Short Self Test (SST)

If the current mode is BiLevel in the PL state and the current breath is spontaneous
and the spontaneous start time has occurred during PH, the time t until the first
VIM of the new A/C mode is the lesser of:

PEEP transition time + 2.5 x duration of the spontaneous inspiration, or

the start time of the spontaneous breath + the length of the apnea interval
(TA).

If the command to change to SIMV occurs after the restricted phase of exhalation has ended, and before a next breath or the apnea interval has elapsed,
the ventilator delivers the first SIMV VIM at the moment the command is recognized.
The point at which the new rate is phased in depends on the current phase of
the SIMV interval and when the rate change command is accepted. If the rate
change occurs during the mandatory interval, the maximum mandatory interval is that for the new or old rate, whichever is less. If the patient generates a
successful inspiratory effort during the spontaneous interval, the ventilator
responds by delivering a spontaneous breath.
Respiratory rate changes are phased in during exhalation only. The new SIMV
interval is determined by the new respiratory rate and is referenced to the start
of the current SIMV period interval, following these rules:

Inspiratory time (TI) of current breath is neither truncated nor extended.

The new inspiration is not delivered until 200 ms of exhalation have elapsed.

The time until the new SIMV interval begins is

10.11

whichever is greater: the new SIMV period interval or 3.5 x the last or current TI,

but not greater than the current SIMV period interval.

Spontaneous (SPONT) Mode


In SPONT mode, the patient initiates inspiration according to the trigger type
in effect, but OIM breaths are allowed which are delivered with the currently
specified mandatory breath parameters. The following spontaneous breath
types are available in SPONT mode:

PS

VS

TC

Operators Manual

10-39

Theory of Operations

PAV+ (if the PAV+ option is installed)

The inspiratory phase begins when the ventilator detects patient effort during
the ventilators exhalation phase. Breath delivery during the inspiratory phase
is determined by the settings for pressure support, PEEP, rise time%, and expiratory sensitivity, unless the breath is an OIM breath.
If Tube Compensation (TC), or Proportional Assist Ventilation (PAV+) (if the
PAV+, option is installed) is selected as the spontaneous type, breath delivery
during the inspiratory phase is determined by the settings for% support (%
Supp), expiratory sensitivity, tube ID, and tube type.

Note:
Given the current ventilator settings, if PAV+ would be an allowable spontaneous type
(except that tube ID < 6 mm) then PAV+ becomes selectable. If selected, tube ID is set
to its New Patient default value based on the PBW entered. An attention icon for tube
ID appears.

If Volume Support (VS) is selected as the spontaneous type, breath delivery


during the inspiratory phase is determined by rise time %, volume support level
(VT SUPP), expiratory sensitivity, and PEEP.
Inspiratory pauses are only possible during OIM breaths, and expiratory pauses
are not allowed during SPONT.
Expiratory trigger methods include:

10.11.1

ESENS (% flow deceleration from peak inspiratory flow)

PBW based time limit (TI too long)

1PPEAK

Inspiratory tidal volume limit (for VS only)

IE Sync (if the IE Sync option is installed)

Airway Pressure Cycling method

Changing to SPONT Mode


If the operator changes to SPONT mode during an A/C or SIMV inspiration
(mandatory or spontaneous), the inspiration is completed, unaffected by the
mode change. Because SPONT mode has no special breath timing require-

10-40

Operators Manual

Short Self Test (SST)

ments, the ventilator then enters the exhalation phase and waits for the detection of patient inspiratory effort, a manual inspiration, or apnea detection.
10.12

Apnea Ventilation
When a patient stops breathing or is no longer being ventilated, it is called
apnea. When apnea is detected by the ventilator the ventilator alarms and
delivers apnea ventilation according to the current apnea ventilation settings.

10.12.1

Apnea Detection
The ventilator declares apnea when no breath has been delivered by the time
the operator-selected apnea interval elapses, plus a small increment of time
(350 ms). This increment allows time for a patient who has begun to initiate a
breath to trigger inspiration and prevent the ventilator from declaring apnea
when the apnea interval is equal to the breath period.
The apnea timer resets whenever an inspiration begins, regardless of whether
the inspiration is patient-triggered, ventilator-triggered, or operator-initiated.
The ventilator then sets a new apnea interval beginning from the start of the
current inspiration. To hold off apnea ventilation, another inspiration must be
delivered before (the current apnea interval + 350 ms) elapses. Apnea detection is suspended during a disconnect, occlusion, or safety valve open (SVO)
state.
Apnea is not declared when the apnea interval setting equals or exceeds the
breath period. For example, if the respiratory rate setting is 4/min, an apnea
interval of 15 s or more means apnea cannot be detected. The ventilator bases
apnea detection on inspiratory (not expiratory) flow, and allows detection of a
disconnect or occlusion during apnea ventilation. Apnea detection is designed
to accommodate interruptions to the typical breathing pattern due to other
ventilator features that temporarily extend the inspiratory or expiratory intervals (rate changes, for example) but still detect a true apnea event.
The following figure shows an apnea breath where TA equals the breath
period.

Operators Manual

10-41

Theory of Operations

Figure10-19.Apnea Interval Equals Breath Period

Tb0

PIM

Tb1

TA (apnea interval)

The figure below shows an apnea breath with TA greater than the breath
period.
Figure10-20.Apnea Interval Greater Than Breath Period

Tb0

VIM

Tb1

TA (apnea interval)

PIM

The following figure shows an apnea breath with TA less than the breath
period.

10-42

Operators Manual

Short Self Test (SST)

Figure10-21.Apnea Interval Less Than Breath Period

Tb0

Apnea interval

Tb1

Apnea Tb0

PIM

Apnea ventilation

Dashed line indicates a PIM to avoid


apnea

Tb (TA<Tb)

Apnea VIM

10.12.2

Transition to Apnea Ventilation


When apnea is declared, the ventilator delivers apnea ventilation according to
the current apnea ventilation settings and displays the apnea settings on the
graphical user interface (GUI). Regardless of the apnea interval setting, apnea
ventilation cannot begin until inspiration of the current breath is complete and
the restricted phase of exhalation has elapsed.

10.12.3

Settings Changes During Apnea Ventilation


All apnea and non-apnea settings remain active on the GUI during apnea ventilation. Both non-apnea and apnea settings changes are phased in according
to the applicable rules. If apnea ventilation is active, new settings are accepted
but not implemented until non-apnea ventilation begins. Allowing key entries
after apnea detection allows adjustment of the apnea interval at setup, regardless of whether apnea has been detected. During apnea ventilation, the
Manual Inspiration key is active, but Expiratory Pause and Inspiratory Pause
keys are not active. The increase O2 control is active during apnea ventilation,
because apnea detection is likely during suctioning.

Operators Manual

10-43

Theory of Operations

The apnea respiratory rate must be 60/TA. Additionally, apnea settings


cannot result in an I:E ratio > 1.00:1.
10.12.4

Resetting Apnea Ventilation


Apnea ventilation is intended as an auxiliary mode of ventilation when there is
insufficient breath delivery to the patient over a specified period of time.
Apnea ventilation can be reset to normal ventilation by the operator (by pressing the Alarm Reset key) or the patient (autoreset). It is also reset when a rate
change is made that renders apnea ventilation inapplicable.
If the patient regains inspiratory control, the ventilator returns to the operatorselected mode of non-apnea ventilation. The ventilator determines whether
the patient has regained respiratory control by monitoring triggered inspirations and exhaled volume. If the patient triggers two consecutive inspirations,
and the exhaled volume is equal to or greater than 50% of the delivered
volume (including any compliance volume), the ventilator resets to non-apnea
ventilation. Exhaled volume is monitored to avoid resetting due to autotriggering caused by large leaks in the patient circuit.

10.12.5

Apnea Ventilation in SIMV


The following strategy is designed to allow SIMV to avoid triggering apnea
ventilation if a VIM breath can be delivered instead:

If the apnea interval (TA) elapses at any time during the mandatory interval, the
ventilator delivers a VIM rather than beginning apnea ventilation.

If TA elapses during the spontaneous interval, apnea ventilation begins.

The following figure shows an illustration of how SIMV is designed to deliver


a VIM rather than trigger apnea ventilation, when possible.

10-44

Operators Manual

Short Self Test (SST)

Figure10-22.Apnea Ventilation in SIMV

10.12.6

Tb

TA

Last breath (PIM)

Tm (If TA elapses during Tm, ventilator delivers a VIM


rather than beginning apnea ventilation

VIM

Ts

Tm max

Phasing in New Apnea Intervals


How a new apnea interval is phased in depends on whether or not apnea ventilation is active. If apnea ventilation is active, the ventilator accepts and implements the new setting immediately. During normal ventilation (that is, apnea
ventilation is not active), these rules apply

10.13

10.13.1

If the new apnea interval setting is shorter than the current (or temporarily extended) apnea interval, the new value is implemented at the next inspiration.

If the new apnea interval setting is longer than the current (or temporarily extended) apnea interval, the old interval is extended to match the new interval immediately.

Detecting Occlusion and Disconnect


Occlusion
The ventilator detects severe patient circuit occlusions in order to protect the
patient from excessive airway pressures, or from receiving little or no gas.
Occlusions require immediate attention to remedy.
The ventilator detects a severe occlusion if

Operators Manual

10-45

Theory of Operations

The inspiratory or expiratory limb of the breathing circuit is partially or completely


occluded (condensate or secretions collected in a gravity-dependent loop, kinked
or crimped tubing, etc.).

The ventilator EXHAUST port is blocked or resistance through the port is too high.

The exhalation valve fails in the closed position (occlusion detection at the From
patient port begins after 195 ms of exhalation has passed.)

The ventilator does not detect a severe occlusion if:

The pressure difference between the inspiratory and the expiratory transducers is
less than or equal to 5 cmH2O.

The exhalation valve fails in the closed position and the pressure in the exhalation
limb is less than 2 cmH2O.

Silicone tubing is attached to the EXHAUST port of the ventilator (i.e. for metabolic monitoring purposes).

The ventilator checks the patient circuit for occlusions during all modes of
breathing (except Stand-by state and Safety Valve Open) at delivery of every
breath. Once the circuit check begins, the ventilator detects a severe occlusion
of the patient circuit within 200 ms. The ventilator checks the EXHAUST port
for occlusions during the expiratory phase of every breath (except during disconnect and safety valve open). Once the EXHAUST port check begins, the
ventilator detects a severe occlusion within 100 ms following the first 200 ms
of exhalation. All occlusion checking is disabled during pressure sensor autozeroing.
When an occlusion is detected, an alarm sounds, the ventilator enters the OSC
(Occlusion Status Cycling) state and displays a message indicating the length
of time the patient has gone without ventilation (how long the ventilator has
been in OSC). This alarm has the capability to autoreset, since occlusions such
as those due to patient activity (for example, crimped, or kinked tubing) can
correct themselves.
Once a severe occlusion is detected, the ventilator acts to minimize airway
pressure. Because any severe occlusion places the patient at risk, the ventilator
minimizes the risk while displaying the length of time the patient has been
without ventilatory support. Severe occlusion is detected regardless of what
mode or triggering strategy is in effect. When a severe occlusion is detected,
the ventilator terminates normal ventilation, terminates any active alarm
silence, annunciates an occlusion alarm, and enters the safe state (exhalation
and inspiratory valve de-energized and safety valve open) for 15 s or until inspiratory pressure drops to 5 cmH2O or less, whichever comes first.

10-46

Operators Manual

Short Self Test (SST)

During a severe occlusion, the ventilator enters OSC, in which it periodically


attempts to deliver a pressure-based breath while monitoring the inspiration
and expiration phases for the existence of a severe occlusion. If the severe
occlusion is corrected, the ventilator detects the corrected condition after two
complete OSC breath periods during which no occlusion is detected. When the
ventilator delivers an OSC breath, it closes the safety valve and waits 500 ms
for the safety valve to close completely, delivers a breath with a target pressure
of 15 cmH2O for 2000 ms, then cycles to exhalation. This breath is followed
by a mandatory breath according to the current settings, but with PEEP = 0 and
O2% equal to 100% for adult/pediatric circuit types or 40% for neonatal circuits. During OSC (and only during OSC), the 2PPEAK (high circuit pressure)
alarm limit is disabled to ensure it does not interfere with the ability of the ventilator to detect a corrected occlusion. When the ventilator does not detect a
severe occlusion, it resets the occlusion alarm, re-establishes PEEP, and reinstates breath delivery according to current settings.
Inspiratory and expiratory pause, and manual inspirations are suspended
during a severe occlusion. Pause maneuvers are canceled by a severe occlusion.
During a severe occlusion, ventilator settings changes are possible. Severe
occlusions are not detected when the ventilator is in the Safety Valve Open
(SVO) state.
A corrected occlusion is detected within 15 s.
10.13.2

Disconnect
A circuit disconnect condition is detected when the ventilator cannot ensure
that a patient is receiving sufficient tidal volume (due to a large leak or disconnected patient circuit). This discussion applies when Leak Sync is disabled.
When a disconnect is detected, an alarm sounds, the ventilator indicates that
a disconnect has been detected, and displays a message indicating the length
of time the patient has gone without ventilation.
Patient data are not displayed during a circuit disconnect condition.
The ventilator monitors the expiratory pressure and flow, delivered volume,
and exhaled volume to declare a disconnect using any of these methods

Operators Manual

The ventilator detects a disconnect when the expiratory pressure transducer measures no circuit pressure and no exhaled flow during the first 200 ms of exhalation. The ventilator postpones declaring a disconnect for another 100 ms to allow
an occlusion (if detected) to be declared first, because it is possible for an occlusion to match the disconnect detection criteria.

10-47

Theory of Operations

Despite many possible variations of circuit disconnections and/or large leaks, it is


possible for a patient to generate some exhaled flow and pressure. The ventilator
then uses the disconnect sensitivity (DSENS, the percentage of delivered volume
lost during the exhalation phase of the same breath to declare a disconnect)
setting to detect a disconnect.

If the disconnect occurs during a spontaneous breath, a disconnect is declared


when the inspiration is terminated by maximum inspiratory time (or the 2TI SPONT
limit setting when Vent Type is non-invasive [NIV]) and the ventilator detects inspiratory flow rising to the maximum allowable.

If the disconnect occurs at the endotracheal tube, the exhaled volume will be
much less than the delivered volume for the previous inspiration. The ventilator
declares a disconnect if the exhaled volume is lower than the DSENS setting for
three consecutive breaths. The DSENS setting helps avoid false detections due to
leaks in the circuit or the patients lungs, and the three-consecutive-breaths
requirement helps avoid false detections due to a patient out-drawing the ventilator during volume control (VC) breaths.

Flow less than a value determined using the DSENS setting and pressure less than
0.5 cmH2O detected for ten (10) consecutive seconds during exhalation.

WARNING:
When vent type is NIV, and DSENS setting is turned OFF, the system may not
detect large leaks and some disconnect conditions it would declare as alarms
during INVASIVE ventilation.

Once the ventilator detects a patient circuit disconnect, the ventilator declares
a high-priority alarm and discontinues breath delivery, regardless of what
mode (including apnea) was active when the disconnect was detected. If there
is an active alarm silence when the disconnect occurs, the alarm silence is NOT
cancelled. The ventilator displays the length of time the patient has been
without ventilatory support. During the disconnect, the exhalation valve closes,
idle flow (10 L/min flow at 100% O2 or 40% O2 in NeoMode, if available with
Leak Sync disabled and 20 L/min with Leak Sync enabled) begins, and breath
triggering is disabled. A message appears identifying how long the patient has
gone without ventilatory support.
The ventilator monitors both expiratory flow and circuit pressures to detect
reconnection. The ventilator declares a reconnect if any of the following criteria are met for the applicable time interval:

10-48

Exhaled idle flow within the reconnect threshold is detected.

Inspiratory and expiratory pressures are both above or both below reconnect
threshold levels or,

Operators Manual

Short Self Test (SST)

Inspiratory pressure rises to a reconnect level.

If the disconnect condition is corrected, the ventilator detects the corrected


condition within one second.
Ventilator triggering, apnea detection, expiratory and inspiratory pause,
manual inspirations, and programmed maneuvers or one-time events are suspended during a patient circuit disconnect condition. Spirometry is not monitored during a disconnect, and all alarms based on spirometry values are
disabled. During a disconnect condition, ventilator settings changes are possible.
If the disconnect alarm is autoreset or manually reset, the ventilator re-establishes PEEP. Once PEEP is reestablished, the ventilator reinstates breath delivery
according to settings in effect before the disconnect was detected.
Circuit disconnect detection is not active during OSC, SVO, or prior to patient
connection.
10.13.3

Annunciating Occlusion and Disconnect Alarms


Occlusion and disconnection cannot be declared at the same time. Therefore,
the ventilator annunciates only the first event to be declared.

10.14

Respiratory Mechanics
Reference Respiratory Mechanics Maneuvers, p. 4-28 for instructions on how
to perform these maneuvers.
In addition to Inspiratory Pause and Expiratory Pause maneuvers, the ventilator
can provide other respiratory maneuvers, including Negative Inspiratory Force
(NIF), Occlusion Pressure (P0.1) and Vital Capacity (VC), as well as automatic calculations of lung function and performance, such as Dynamic Compliance
(CDYN) and Dynamic Resistance (RDYN), Peak Expiratory Flow (PEF), End Expiratory Flow (EEF), C20/C, and Peak Spontaneous Flow (PSF).
Respiratory maneuvers can be performed in all breathing modes (except as
noted below) but are not available during the following conditions:

Apnea ventilation

Safety PCV

Occlusion Status Cycling (OSC)

Operators Manual

10-49

Theory of Operations

Non-invasive ventilation (NIV)

When the circuit type is neonatal

SVO

Ventilator is in Stand-by state

When any other respiratory maneuver has already taken place during the same
breath

The GUI also displays any maneuver request, distinguishing between requests
that are accepted or rejected, and any maneuver that has begun, ended, or has
been canceled.
When a maneuver is selected, a GUI information panel is opened, displaying
the maneuver name, user prompts and controls, and recent calculated results.
Any maneuver is canceled automatically upon declaration of any of the following alarms:

1PPEAK alarm

1PVENT alarm

1VTI

The following Respiratory Mechanics maneuvers are not available in BiLevel


ventilation:

10.14.1

Occlusion Pressure

NIF Negative Inspiratory Force

VC Vital Capacity

Inspiratory Pause
Note:
Inspiratory pause and expiratory pause maneuvers can be performed directly by
pressing the respective keys on the GUI or by swiping the Menu tab on the left side
of the GUI. For more information on how to perform Respiratory Mechanics
Maneuvers from the Menu tab, Reference Respiratory Mechanics Maneuvers, p. 4-28.

An inspiratory pause extends the inspiratory phase of a single mandatory


breath for the purpose of measuring end inspiratory circuit pressure which is
used to calculate static compliance of the patients lungs and thorax (CSTAT),

10-50

Operators Manual

Short Self Test (SST)

static resistance of the respiratory system (RSTAT), and inspiratory plateau pressure (PPL). To calculate these pressures, the inspiratory and exhalation valves
are closed, allowing pressures on both sides of the artificial airway to equalize,
revealing the actual lung inflation pressure during a no-flow condition. An
inspiratory pause can be either automatically or manually administered, and is
only available during the next mandatory breath in A/C, SIMV, BiLevel or
SPONT modes. In BiLevel, an inspiratory pause maneuver is scheduled for the
next inspiration prior to a transition from PH to PL. Only one inspiratory pause
is allowed per breath. An inspiratory pause cannot occur during apnea ventilation, safety PCV, Stand-by state, Occlusion, and SVO.
An automatic inspiratory pause begins when the inspiratory pause key is
pressed momentarily or the maneuver is started from the GUI screen. Reference To access respiratory mechanics maneuvers, p. 4-28 for more information
on performing respiratory mechanics maneuvers from the Menu tab on the
GUI rather than using the keys on the GUI. The pause lasts at least 0.5 second
but no longer than three(3) s. A manual inspiratory pause starts by pressing
and holding the inspiratory pause key. The pause lasts for the duration of the
key press (up to seven(7) s).
An active manual inspiratory pause is considered complete if any of the following occur:

The inspiratory pause key is released and at least two seconds of inspiratory pause
have elapsed or pressure stability conditions have been detected for not less than
0.5 s.

Pause duration reaches seven (7) s.

A manual inspiratory pause maneuver request (if the maneuver is not yet
active) will be canceled if any of events1-10 occur. Reference Inspiratory and
Expiratory Pause Events.
Table10-5.Inspiratory and Expiratory Pause Events
Event Identifier

Operators Manual

Event

There is a loss of communications with the GUI

High ventilator pressure limit (2PVENT) is reached

High circuit pressure limit (2PPEAK) is reached

A disconnect is detected

Occlusion is detected

10-51

Theory of Operations

Table10-5.Inspiratory and Expiratory Pause Events (Continued)


Event Identifier

Event

Apnea is detected

72 seconds have elapsed without an inspiratory


pause after one has been requested

INSPIRATION TOO LONG alarm is detected

High inspired tidal volume (1VTI) alarm is detected

10

High compensation pressure (1PCOMP) alarm is


detected

11

Cancel is touched if maneuver is initiated from the


GUI screen.

12

Safety Valve Open (SVO) is detected

13

Patient trigger effort causes circuit pressure to go


below sensitivity. The sensitivity level is the setting
value for pressure trigger or the backup pressure
value for flow trigger

14

BUV is entered

15

Expiratory pause key is pressed (Inspiratory pause key


if maneuver is an expiratory pause)

During a manual inspiratory pause, the maneuver is terminated if any of


events 1,3,5,6,12,or 13 occur. Reference Inspiratory and Expiratory Pause
Events, p. 10-51.
An inspiratory pause maneuver is ignored if the ventilator is in Apnea ventilation, safety PCV, OSC, SVO, BUV, or Stand-by state.
An active automatic inspiratory pause maneuver is terminated and exhalation
begun if any of events1-12, or 14 occur. Reference Inspiratory and Expiratory
Pause Events, p. 10-51.
The active automatic inspiratory pause maneuver is considered complete if
the pause duration reaches three seconds or pressure stability conditions have
been detected for not less than 0.5 s.
An automatic inspiratory pause maneuver request (if the maneuver is not yet
active) will be canceled if any of events1-9, 11, 12, 14, or 15 occur. Reference
Inspiratory and Expiratory Pause Events, p. 10-51.
Other characteristics of inspiratory pause include:

10-52

Operators Manual

Short Self Test (SST)

During an inspiratory pause, the apnea interval (TA) is extended by the duration of
the inspiratory pause.

If the ventilator is in SIMV, the breath period during which the next scheduled VIM
occurs will also be extended by the amount of time the inspiratory pause is active.

All activations of the inspiratory pause control are logged in the Patient Data Log.

Severe occlusion detection is suspended

When calculating I:E ratio, inspiratory pause is considered part of the inspiration
phase.

The expiratory time remains unchanged, and will result in a change in the I:E ratio
for the breath that includes the inspiratory phase.

Once the inspiratory Pause maneuver is completed the operator can review the
quality of the maneuver waveform and accept or reject the maneuver data.
10.14.2

Expiratory Pause
An expiratory pause extends the exhalation phase of a single breath in order
to measure end expiratory lung pressure (PEEPTOT) and allows intrinsic PEEP
(PEEPI) to be calculated as PEEPTOT minus set PEEP. The pressures on either side
of the artificial airway are allowed to equalize by closing the inspiratory and
exhalation valves. Expiratory pause is available in A/C, SIMV, and BiLevel
modes. For A/C and SIMV, the expiratory pause maneuver is scheduled for the
next end-of-exhalation prior to a mandatory breath. In BiLevel, the expiratory
pause occurs at the next end-of-exhalation prior to a transition from PL to PH.
Only one expiratory pause per breath is allowed, and the expiratory pause
request is rejected if an inspiratory pause has already taken place during the
same breath.
A request for an expiratory pause maneuver is ignored in apnea ventilation,
safety PCV, SPONT, OSC, BUV, and Stand-by. Reference To access respiratory
mechanics maneuvers, p. 4-28 for more information on performing these
maneuvers from the GUI screen rather than using the keys on the GUI.
Either manual or automatic expiratory pause maneuvers can occur. A
momentary press of the expiratory pause key begins an automatic expiratory
pause which lasts at least 0.5 s, but no longer than 3.0 s. A manual expiratory
pause starts by pressing and holding the expiratory pause key and lasts for the
duration of the key-press (up to 15 s).

Operators Manual

10-53

Theory of Operations

An active manual expiratory pause is terminated if any of events1-12 occur.


Reference Reference Inspiratory and Expiratory Pause Events, p. 10-51.
An active manual expiratory pause is complete if the expiratory pause key is
released and at least three (3) s of expiratory pause have elapsed, pressure stability conditions have been detected for 0.5 s, or pause duration lasts 15 s.
An active automatic expiratory pause is terminated if any of events1,3, or 1113 occur. Reference Inspiratory and Expiratory Pause Events, p. 10-51.
An active automatic expiratory pause is complete if pause duration reaches
three (3) s or pressure stability conditions have been detected for 0.5 s, or
pause duration lasts 15 s.
The automatic expiratory pause maneuver request (the maneuver is not yet
active) is canceled if events1-9, 11, 12, or 15 occur. Reference Inspiratory and
Expiratory Pause Events, p. 10-51.
The automatic expiratory pause maneuver is terminated and inspiration
begun if any of events1,3,or 11-13 occur. Reference Inspiratory and Expiratory
Pause Events, p. 10-51.
Other characteristics of expiratory pause include:

During an active manual expiratory pause, severe occlusion detection is suspended.

When calculating I:E ratio, the expiratory pause is considered part of the exhalation phase.

During the expiratory pause, the inspiratory time remains unchanged, so the I:E
ratio is changed for the breath that includes the expiratory pause.

All activations of the expiratory pause control are logged in the Patient Data log.

Once the expiratory pause maneuver is completed the operator can review the
quality of the maneuver waveform and accept or reject the maneuver data.
10.14.3

Negative Inspiratory Force (NIF) Maneuver


The Negative Inspiratory Force (NIF) maneuver is a coached maneuver where
the patient is prompted to draw a maximum inspiration against an occluded
airway (the inspiratory and exhalation valves are fully closed).
A NIF maneuver is canceled if:

10-54

Disconnect is detected

Occlusion is detected

Operators Manual

Short Self Test (SST)

SVO is detected

1PPEAK alarm is declared

1PVENT alarm is declared

1VTI alarm is declared

Communications with the GUI is lost

The maneuver has been active for 30 s and an inspiration is not detected

INSPIRATION TOO LONG alarm is declared

A manual inspiration is requested

When a NIF maneuver is activated, a single pressure-time waveform grid is


automatically displayed. During a NIF maneuver, the circuit pressure displays
on the waveforms screen and is regularly updated, producing a real-time display.
When an active NIF maneuver ends successfully, the calculated NIF result
appears on the waveforms screen and on the maneuver panel. The NIF value
displayed represents the maximum negative pressure from PEEP.
When a NIF maneuver ends, a PEEP restoration breath is delivered to the
patient, then normal breath delivery resumes.
10.14.4

P0.1 Maneuver (Occlusion Pressure)


P0.1 is the negative airway pressure (delta pressure change) generated during
the first 100 ms of an occluded inspiration. It is an estimate of the neuromuscular drive to breathe.
When a P0.1 maneuver ends successfully, the calculated airway pressure displays on the waveforms screen and on the maneuver panel. A P0.1 maneuver
is terminated if seven (7) s elapse and a trigger has not been detected to activate the maneuver.
A P0.1 maneuver is canceled if

Disconnect is detected

Occlusion is detected

SVO is detected

Operators Manual

10-55

Theory of Operations

10.14.5

1PPEAK alarm is declared

1PVENT alarm is declared

1VTI alarm is declared

INSPIRATION TOO LONG alarm is declared

Communications with the GUI is lost

A manual inspiration is requested

Vital Capacity (VC) Maneuver


The Vital Capacity (VC) maneuver is a coached maneuver where the patient is
prompted to draw a maximum inspiration (regardless of the current settings)
and then to slowly and fully exhale.
When the Vital Capacity maneuver becomes active, the ventilator delivers a
spontaneous inspiration in response to patient effort (with
PSUPP = 0, Rise time % = 50, and ESENS = 0), and then allows for a full exhalation effort.
When a Vital Capacity maneuver is requested, a single Volume-Time waveform
grid is automatically displayed. A Vital Capacity maneuver is canceled if:

10-56

Disconnect is detected

Occlusion is detected

SVO is detected

1PPEAK alarm is declared

1PVENT alarm is declared

1VTI alarm is declared

INSPIRATION TOO LONG alarm is declared

Communications with the GUI is lost

A manual inspiration is requested

The maneuver as been active for 15 s and inspiration is not detected

The Cancel is touched

Operators Manual

Short Self Test (SST)

When an active VC maneuver ends successfully, the calculated expiratory


volume displays on the waveforms screen and on the maneuver panel and a
PEEP restoration breath is delivered.
10.15

10.15.1

Ventilator Settings
Apnea Ventilation
Apnea ventilation is a backup mode and starts if the patient fails to breathe
within the apnea interval (TA) set by the operator. TA defines the maximum
allowable length of time between the start of inspiration and the start of the
next inspiration. Available settings include mandatory type (PC or VC). For PC
breaths the allowable settings are

Apnea interval (TA)

Inspiratory pressure (PI)

Inspiratory time (TI)

Respiratory rate (f)

For VC breaths, the allowable settings are:

Apnea interval (TA)

Flow pattern

O2 %

Peak inspiratory flow (VMAX)

Respiratory rate (f)

Tidal Volume (VT)

During apnea ventilation with PC selected as the mandatory type, rise time %
is fixed at 50%, and the constant parameter during a rate change is inspiratory
time (TI).
If apnea is possible (that is, if (60/f) > TA) increasing the non-apnea O2%
setting automatically changes apnea ventilation O2% if it is not already set
higher than the new non-apnea O2%. Apnea ventilation O2% does not automatically change by decreasing the non-apnea O2%. Whenever there is an

Operators Manual

10-57

Theory of Operations

automatic change to an apnea setting, a message appears on the GUI, and the
apnea settings screen appears.
During apnea ventilation, changes to all non-apnea ventilation settings are
allowed, but the new settings do not take effect until the ventilator resumes
normal ventilation. Being able to change TA during apnea ventilation can avoid
immediately re-entering apnea ventilation once normal ventilation resumes.
Because the minimum value for TA is ten (10) s, apnea ventilation cannot take
place when non-apnea f is greater than or equal to 5.8/min.The ventilator does
not enter apnea ventilation if TA is equal to the breath period interval. Set TA
to a value less than the expected or current breath period interval as a way of
allowing the patient to initiate breaths while protecting the patient from the
consequences of apnea.
10.15.2

Circuit Type and Predicted Body Weight (PBW)


Together, circuit type and PBW (displayed in lb or kg) provide the basis for new
patient values and absolute limits on various ventilator settings such as tidal
volume (VT) and Peak flow (VMAX). Run SST in order to change the circuit
type.The table below gives the minimum, maximum, and new patient default
values for VT based on circuit type.
Table10-6.Values for VT Based on Circuit Type
Maximum VT

Circuit Type

New Patient Default

Minimum VT

Neonatal

When mandatory type is


VC+, MAX {2 mL, (mL/kg
Ratio x PBW)} mL;
When mandatory type is
VC, MAX {3 mL, (mL/kg
Ratio * PBW)} mL

2 mL if NeoMode 2.0
software option is
installed

315 mL

Pediatric

mL/kg ratio x PBW mL

25 mL

1590 mL

Adult

mL/kg ratio x PBW mL

25 mL

2500 mL

Reference Ventilator Settings Range and Resolution, p. 11-8, VT setting, for


more information on VT calculations based on PBW and circuit type.

10-58

Operators Manual

Short Self Test (SST)

Table10-7.Peak Flow and Circuit Type (Leak Sync Disabled)


Circuit Type

Maximum peak flow


(VMAX) setting

Neonatal

30 L/min

Pediatric

60 L/min

Adult

150 L/min

PBW determines constants for breath delivery algorithms, some user-settable


alarms, the high spontaneous inspiratory time limit setting
(2TI SPONT) in NIV, and the non-settable INSPIRATION TOO LONG alarm.
10.15.3

Vent Type
There are two Vent Type choices INVASIVE and NIV (non-invasive). INVASIVE
ventilation is conventional ventilation used with endotracheal or tracheostomy
tubes. All installed software options, breathing modes, breath types, and
trigger types are available during INVASIVE ventilation.
NIV interfaces include non-vented full-faced or nasal masks or nasal prongs.
Reference NIV Breathing Interfaces, p. 4-22 for a list of interfaces that have
been successfully tested with NIV).
NIV enables the ventilator to handle large system leaks associated with these
interfaces by providing pressure-based disconnect alarms, minimizing false disconnect alarms, and replacing the INSPIRATION TOO LONG alarm with a High
Spontaneous Inspiratory Time limit (2TI SPONT) setting and visual indicator.
The following list shows the subset of INVASIVE settings active during NIV:

Mode A/C, SIMV, SPONT. (BiLevel is not available during NIV.)

Mandatory Type PC or VC. (VC+ is not available during NIV.)

Spontaneous Type PS (TC and VS are not available during NIV.)

During NIV alarm setup, the clinician may set alarms to OFF and must determine if doing so is appropriate for the patients condition.

Operators Manual

10-59

Theory of Operations

10.15.4

Mode and Breath Type


Specifying the mode defines the types and sequences of breaths allowed for
both INVASIVE and NIV Vent Types.
Table10-8.Modes and Breath Types

Mode

Mandatory Breath
Type

A/C

INVASIVE: VC, VC+, or


PC

Spontaneous Breath
Type

Sequence

Not allowed

All mandatory (patient-, ventilator-,


or operator-initiated)

NIV: VC or PC
SIMV

INVASIVE: PC, VC, or


VC+
NIV: VC or PC

Pressure supported (PS) or


TC.

Each new breath begins with a mandatory interval, during which a


patient effort yields a synchronized
mandatory breath. If no patient effort
is detected during the mandatory
interval, the ventilator delivers a mandatory breath. Subsequent patient
efforts before the end of the breath
yield spontaneous breaths.

SPONT

Not allowed (PC or VC


allowed only for
manual inspirations).

INVASIVE: Pressure supported (PS), Tube compensated


(TC), Volume supported
(VS), Proportionally assisted
(PAV+)
NIV: PS

All spontaneous (except for manual


inspirations).

PC

PS, TC

Combines mandatory and spontaneous breathing modes. Reference


Appendix A for more information on
BiLevel ventilation.

VC or PC (allowed only
for OIM breaths)

N/A

All spontaneous (except for manual


inspirations).Reference Appendix E,
NeoMode 2.0 for more information
on CPAP.

BiLevel
(INVASIVE
vent type
only)
CPAP

Breath types must be defined before settings can be specified. There are only
two categories of breath type: mandatory and spontaneous. Mandatory
breaths are volume controlled (VC) or pressure controlled (PC or VC+). The
ventilator currently offers spontaneous breaths that are pressure supported
(PS) volume supported (VS), tube compensated (TC), or proportionally assisted
(PAV+), if the PAV+ option is installed. The figure below shows the modes and
breath types available on the ventilator.

10-60

Operators Manual

Short Self Test (SST)

A/C,
SIMV,
SPONT,
BiLevel
Mandatory
PC

VC

Spontaneous
VC+

PS

TC

VS

PAV+

The mode setting defines the interaction between the ventilator and the
patient.

Assist/control (A/C) mode allows the ventilator to control ventilation within


boundaries specified by the practitioner. All breaths are mandatory, and can be
PC, VC, or VC+.

Spontaneous (SPONT) mode allows the patient to control ventilation. The patient
must be able to breathe independently, and exert the effort to trigger ventilator
support.

Synchronous Intermittent Mandatory Ventilation (SIMV) is a mixed mode that


allows a combination of mandatory and spontaneous interactions. In SIMV, the
breaths can be spontaneous or mandatory, mandatory breaths are synchronized
with the patient's inspiratory efforts, and breath delivery is determined by the f
setting.

BiLevel is a mixed mode that combines both mandatory and spontaneous breath
types. Breaths are delivered in a manner similar to SIMV mode with PC selected,
but providing two levels of pressure. The patient is free to initiate spontaneous
breaths at either pressure level during BiLevel.

Changes to the mode are phased in at the start of inspiration. Mandatory and
spontaneous breaths can be flow- or pressure-triggered, or IE Sync triggered
(if the IE Sync software option is installed).
The ventilator automatically links the mandatory type setting to the mode setting. During A/C or SIMV modes, once the operator has specified volume or
pressure, the ventilator displays the appropriate breath parameters. Changes
in the mandatory type are phased in at the start of inspiration.
10.15.5

Respiratory Rate (f)


The f setting determines the minimum number of mandatory breaths per
minute for ventilator-initiated mandatory breaths in A/C, SIMV, and BiLevel
modes.If the mode is A/C or SIMV and VC is the breath type, specifying VMAX
and flow pattern determines TI, TE, and I:E. In PC breaths, specifying TI auto-

Operators Manual

10-61

Theory of Operations

matically determines the other timing variables. Reference Inspiratory Time (TI),
p. 10-65 for an explanation of the interdependencies of f, TI, TE, and I:E.
Changes to the f setting are phased in at the start of inspiration.
The ventilator does not accept a proposed f setting if it would cause the new
TI or TE to be less than 0.2 second, the TI to be greater than eight(8) s, or I:E
ratio greater than 4.00:1. (The ventilator also applies these restrictions to a proposed change to the apnea respiratory rate, except that apnea I:E cannot
exceed 1.00:1. An exception to this rule occurs in BiLevel ventilation where the
proposed f setting will allow the I:E ratio to be greater than 4.00:1 only until
the minimum TL is reached.
10.15.6

Tidal Volume (VT)


The VT setting determines the volume of gas delivered to the patient during a
VC mandatory breath. The delivered VT is compensated for BTPS and patient
circuit compliance. Changes to the VT setting are phased in at the start of inspiration. The VT setting only affects the delivery of mandatory breaths.
When proposing a change to the VT setting, the ventilator compares the new
value with the settings for f, VMAX, flow pattern, and TPL. If the proposed
setting would result in an I:E ratio that exceeds 4.00:1 or a TI greater than
eight(8) s or less than 0.2 s, or a TE less than 0.2 s, the ventilator disallows the
change.

10.15.7

Peak Inspiratory Flow (VMAX)


The VMAX setting determines the maximum rate of delivery of tidal volume to
the patient during mandatory VC breaths, only. Changes to VMAX are phased
in at the start of inspiration. Mandatory breaths are compliance compensated,
even at the maximum VMAX setting. Circuit compliance compensation does
not cause the ventilator to exceed the ventilators maximum flow capability.
When proposing a change to the VMAX setting, the ventilator compares the
new value with the settings for VT, f, flow pattern, and TPL. It is impossible to
set a new VMAX that would result in an I:E ratio that exceeds 4.00:1, or a TI
greater than 8.0 s or less than 0.2 s, or a TE less than 0.2 s.

10-62

Operators Manual

Short Self Test (SST)

10.15.8

Plateau Time (TPL)


The TPL setting determines the amount of time inspiration is held in the
patient's airway after inspiratory flow has ceased. TPL is available only during
VC mandatory breaths (for A/C and SIMV mode, and operator-initiated mandatory breaths). TPL is not available for PC mandatory breaths. Changes to the
TPL setting are phased in at the start of inspiration.
When proposing a change to the TPL setting, the ventilator computes the new
I:E ratio and TI, given the current settings for VT, f, VMAX, and flow pattern. It
is impossible to set a new TPL that would result in an I:E ratio that exceeds
4.00:1, or a TI greater than eight(8) s or less than 0.2 s, or a TE less than 0.2 s.
For the I:E ratio calculation, TPL is considered part of the inspiration phase.

10.15.9

Flow Pattern
The flow pattern setting defines the gas flow pattern of volume-controlled
(VC) mandatory breaths only. The selected values for VT and VMAX apply to
both the square or descending ramp flow patterns. If VT and VMAX and are
held constant, TI approximately halves when the flow pattern changes from
descending ramp to square (and approximately doubles when flow pattern
changes from square to descending ramp), and corresponding changes to the
I:E ratio also occur. Changes in flow pattern are phased in at the start of inspiration.
The settings for flow pattern, VT,f, TPL, and VMAX are interrelated. If any setting
change would cause any of the following, the ventilator does not allow that
change

I:E ratio > 4:1

TI > 8.0 s or TI < 0.2 s

TE < 0.2 s

10.15.10 Flow

Sensitivity (VSENS)

The VSENS setting defines the rate of flow inspired by a patient that triggers the
ventilator to deliver a mandatory or spontaneous breath. When V-TRIG is selected, a base flow of gas (1.5 L/min) travels through the patient circuit during the

Operators Manual

10-63

Theory of Operations

ventilators expiratory phase. Once a value for flow sensitivity is selected, the
ventilator delivers a base flow equal to VSENS + 1.5 L/min (base flow is not userselectable). When the patient inhales and their inspiratory flow exceeds the
VSENS setting, a trigger occurs and the ventilator delivers a breath. Reductions
to VSENS are phased in immediately, while increases are phased in at the start
of exhalation.
When VSENS is active, it replaces pressure sensitivity (PSENS). The VSENS setting
has no effect on the PSENS setting. VSENS can be active in any ventilation mode
(including pressure supported, volume controlled, pressure controlled, and
apnea ventilation). When VSENS is active, a backup PSENS setting of 2 cmH2O is
in effect to detect the patient's inspiratory effort, even if the flow sensors do
not detect flow.
Although the minimum VSENS setting of 0.2 L/min (adult/pediatric circuit types)
or 0.1 L/min (neonatal circuit type) can result in autotriggering, it can be appropriate for very weak patients. The maximum setting of 20 L/min (adult/pediatric
circuit types) or 10 L/min (neonatal circuit type) is intended to avoid autotriggering when there are significant leaks in the patient circuit.
10.15.11 Pressure

Sensitivity (PSENS)

The PSENS setting selects the pressure drop below baseline (PEEP) required to
begin a patient-initiated breath (either mandatory or spontaneous). Changes
to PSENS are phased in immediately. The PSENS setting has no effect on the
VSENS setting and is active only if the trigger type is P-TRIG.
Lower PSENS settings provide greater patient comfort and require less patient
effort to initiate a breath. However, fluctuations in system pressure can cause
autotriggering at very low settings. The maximum PSENS setting avoids autotriggering under worst-case conditions if patient circuit leakage is within specified limits.
10.15.12 Inspiratory

Pressure (PI)

The PI setting determines the pressure at which the ventilator delivers gas to
the patient during a PC mandatory breath. The PI setting only affects the delivery of PC mandatory breaths. The selected PI is the pressure above PEEP. (For
example, if PEEP is set to five cmH2O, and PI is 20 cmH2O, the ventilator deliv-

10-64

Operators Manual

Short Self Test (SST)

ers gas to the patient at 25 cmH2O.) Changes to the PI setting are phased in
at the start of inspiration.
The sum of PEEP + PI + 2 cmH2O cannot exceed the high circuit pressure
(2PPEAK) limit. To increase this sum of pressures, first raise the 2PPEAK limit
before increasing the settings for PEEP or PI. The minimum value for PI is 5
cmH2O and the maximum value is 90 cmH2O.
10.15.13 Inspiratory

Time (TI)

The TI setting determines the time during which an inspiration is delivered to


the patient for PC mandatory breaths. The ventilator accepts a setting as long
as the resulting I:E ratio and TE settings are valid. Changes to TI phase in at the
start of inspiration. Directly setting TI in VC mandatory breaths is not allowed.
The ventilator rejects settings that result in an I:E ratio greater than 4.00:1, a
TI greater than eight(8) s or less than 0.2 s, or a TE less than 0.2 s to ensure the
patient has adequate time for exhalation.
Setting f and TI automatically determines the value for I:E and TE.
60 f T I = T E
This equation summarizes the relationship between TI, I:E,TE, and breath
period time

60
T I = ------ I :E 1 + I :E
f
If the f setting remains constant, any one of the three variables (TI, I:E, or TE)
can define the inspiratory and expiratory intervals. If the f setting is low (and
additional spontaneous patient efforts are expected), TI can be a more useful
variable to set than I:E. As the f setting increases (and the fewer patient-triggered breaths are expected), the I:E setting becomes more relevant. Regardless
of which variable is chosen, a breath timing bar always shows the interrelationship between TI, I:E,TE, and f.

Operators Manual

10-65

Theory of Operations

10.15.14 Expiratory

Time (TE)

The TE setting defines the duration of exhalation for PC and VC+ mandatory
breaths, only. Changes to the TE setting are phased in at the start of exhalation. Setting f and TE automatically determines the value for I:E ratio and TI.
Reference Inspiratory Time (TI), p. 10-65 for an explanation of the interdependencies of f, TI, TE, and I:E.
10.15.15 I:E

Ratio

The I:E ratio setting is available when I:E is selected as the constant during rate
change. The I:E setting determines the ratio of inspiratory time to expiratory
time for mandatory PC breaths. The ventilator accepts the specified range of
direct I:E ratio settings as long as the resulting TI and TE settings are within the
ranges established for mandatory breaths. Changes to the I:E ratio phase in at
the start of inspiration. Directly setting the I:E ratio in VC mandatory breaths is
not allowed. Reference Inspiratory Time (TI), p. 10-65 for an explanation of the
interdependencies of f, TI, TE, and I:E.
Setting f and I:E automatically determine the values for TI and TE. The
maximum I:E ratio setting of 4.00:1 is the maximum that allows adequate time
for exhalation and is intended for inverse ratio pressure control ventilation.
10.15.16 High

Pressure (PH) in BiLevel

The pressure level entered by the operator for the inspiratory phase of the
mandatory breath in BiLevel ventilation.
10.15.17 Low

Pressure (PL) in BiLevel

The pressure level entered by the operator for the expiratory phase of the mandatory breath in BiLevel ventilation.
10.15.18 High

Time (TH) in BiLevel

The duration of time (in seconds) the ventilator maintains the set high pressure
level in BiLevel ventilation.

10-66

Operators Manual

Short Self Test (SST)

10.15.19 Low

Time (TL) in BiLevel

The duration of time (in seconds) the ventilator maintains the set low pressure
level in BiLevel ventilation.
10.15.20 TH:TL

Ratio in BiLevel

The ratio of TH to TL in BiLevel ventilation, similar to I:E ratio when ventilating


a patient without BiLevel.
10.15.21 PEEP

This setting defines the positive end-expiratory pressure (PEEP), also called
baseline airway pressure. PEEP is the positive pressure maintained in the
patient circuit during exhalation. Changes to the PEEP setting are phased in at
the start of exhalation.
The sum of

PEEP + 7 cmH2O, or

PEEP + PI + 2 cmH2O (if PC is active), or

PEEP + PSUPP + cmH2O (if PS is on)

cannot exceed the 2PPEAK limit. To increase the sum of pressures, first raise the
2PPEAK limit before increasing the settings for PEEP, PI, or PSUPP.
If there is a loss of PEEP from occlusion, disconnect, Safety Valve Open, or loss
of power conditions, PEEP is re-established (when the condition is corrected)
by the ventilator delivering a PEEP restoration breath. The PEEP restoration
breath is a 1.5 cmH2O pressure-supported breath with exhalation sensitivity of
25%, and rise time % of 50%. A PEEP restoration breath is also delivered at
the conclusion of Vent Startup. After PEEP is restored, the ventilator resumes
breath delivery at the current settings.

Note:

PEEP restoration breath parameters are not user adjustable.

Operators Manual

10-67

Theory of Operations

10.15.22 Pressure

Support (PSUPP)

The PSUPP setting determines the level of positive pressure above PEEP applied
to the patient's airway during a spontaneous breath. PSUPP is only available in
SIMV, SPONT, and BiLevel, in which spontaneous breaths are allowed. The
PSUPP setting is maintained as long as the patient inspires, and patient demand
determines the flow rate. Changes to the PSUPP setting are phased in at the
start of inspiration. The pressure support setting affects only spontaneous
breaths.
The sum of PEEP +PSUPP +2 cmH2O cannot exceed the 2PPEAK limit. To increase
the sum of pressures, first raise the 2PPEAK limit before increasing the settings
for PEEP or PSUPP. Since the 2PPEAK limit is the highest pressure considered safe
for the patient, a PSUPP setting that would cause a 1PPEAK alarm requires reevaluating the maximum safe circuit pressure.
10.15.23 Volume

Support (VT SUPP)

Volume support (VT SUPP) is defined as the volume of gas delivered to the
patient during spontaneous VS breaths. Changes to the to the VT SUPP setting
are phased in at the start of inspiration.
10.15.24 %

Supp in TC

In TC, the% Supp setting represents the amount of the imposed resistance of
the artificial airway the TC breath type will eliminate by applying added pressure at the patient circuit wye. For example, if the % Supp setting is 100%, TC
eliminates 100% of the extra work imposed the by the airway. At 50%, TC
eliminates 50% of the added work from the airway. TC is also used with BiLevel, and is available during both PH and PL phases.
10.15.25 %

Supp in PAV+

In PAV+, the % Supp setting represents the percentage of the total work of
breathing provided (WOB) by the ventilator. Higher inspiratory demand yields
greater support from the ventilator. The patient performs the remaining work.
If the total WOB changes (resulting from a change to resistance or compliance)
the percent support remains constant.

10-68

Operators Manual

Short Self Test (SST)

10.15.26 Rise

Time %

The rise time % setting allows adjustment of the speed at which the inspiratory
pressure reaches 95% of the target pressure. Rise time settings apply to PS
(including a setting of 0 cmH2O),VS, PC, or VC+ breaths. The higher the value
of rise time %, the more aggressive (and hence, the more rapid) the rise of
inspiratory pressure to the target (which equals PEEP + PI (or PSUPP)). The rise
time % setting only appears when pressure-based breaths are available. The
range of rise time % is 1% to 100%. A setting of 50% takes approximately
half the time to reach 95% of the target pressure as a setting of 1.

For mandatory PC, VC+, or BiLevel breaths, a rise time setting of 1 produces a
pressure trajectory reaching 95% of the inspiratory target pressure (PEEP + PI) in
two (2) s or 2/3 of the TI, whichever is shortest.

For spontaneous breaths (VS, or PS), a rise time setting of 1 produces a pressure
trajectory reaching 95% of the inspiratory target (PEEP + PSUPP) in (0.4 x PBWbased TI TOO LONG x 2/3) s.

When both PC and PS breaths are active, the slopes and thus the pressure trajectories can appear to be different. Changes to TI and PI cause PC pressure trajectories to change. Changes in rise time % are phased in at the start of inspiration.

When PSUPP = 0, the rise time % setting determines how quickly the ventilator
drives circuit pressure to PEEP + 1.5 cmH2O.

10.15.27 Expiratory

Sensitivity (ESENS)

The ESENS setting defines the percentage of the measured peak inspiratory
flow at which the ventilator cycles from inspiration to exhalation in all spontaneous breath types unless IE Sync has been selected for inspiratory and expiratory detection. When inspiratory flow falls to the level defined by ESENS,
exhalation begins. ESENS is a primary setting and is accessible from the GUI
screen. Changes to ESENS are phased in at the next patient-initiated spontaneous inspiration.
ESENS complements rise time %. Rise time % should be adjusted first to match
the patient's inspiratory drive, and then the ESENS setting should cause ventilator exhalation to occur at a point most appropriate for the patient. The higher
the ESENS setting, the shorter the inspiratory time. Generally, the most appropriate ESENS is compatible with the patient's condition, neither extending nor
shortening the patient's intrinsic inspiratory phase.

Operators Manual

10-69

Theory of Operations

ESENS in a PAV+ breath is expressed in L/min instead of percent.


10.15.28 Disconnect

Sensitivity (DSENS)

Disconnect sensitivity (DSENS) is defined as the percentage of returned volume


lost due to a leak, above which the ventilator declares a CIRCUIT DISCONNECT
alarm (when Leak Sync is not enabled). When DSENS is set to its lowest value
(20%) it has the highest sensitivity for detecting a leak or disconnect. Conversely, when DSENS is set to its highest value (95%), the ventilator is least sensitive to a leak or disconnect, because greater than 95% of the returned
volume must be lost before the alarm annunciates. During NIV, the DSENS value
is automatically set to OFF, which is means that returned volume loss is not
considered and the alarm will not sound.DSENS with Leak Sync enabled is
expressed in L/min instead of percent.

Note:
If DSENS is set to OFF during NIV, the ventilator is still capable of declaring a CIRCUIT
DISCONNECT alarm.
Note:
DSENS cannot be turned OFF if Leak Sync is enabled.

Changes to DSENS are phased in at the start of inspiration.


10.15.29 High

Spontaneous Inspiratory Time Limit (2TI SPONT)

The high spontaneous inspiratory time limit setting (2TI SPONT) is available only
in SIMV or SPONT modes during NIV, and provides a means for setting a
maximum inspiratory time after which the ventilator automatically transitions
to exhalation. The default 2TI SPONT setting is based upon circuit type and PBW.
For pediatric/adult circuit types, the new patient default value is (1.99 + (0.02
x PBW)) s
For neonatal circuit types, the new patient default value is ((1.00 + (0.10 x
PBW) s
The 1TI SPONT indicator appears on the primary display at the beginning of a
ventilator-initiated exhalation and remains visible for as long as the ventilator

10-70

Operators Manual

Short Self Test (SST)

truncates breaths in response to the 2TI SPONT setting. The 1TI SPONT indicator
disappears when the patients inspiratory time returns to less than the 2TI
SPONT setting, or after 15 seconds has elapsed after the beginning of exhalation
of the last truncated breath. Changes to 2TI SPONT are phased in at the start of
inspiration.
10.15.30 Humidification

Type

The humidification type setting sets the type of humidification system (heated
expiratory tube, non-heated expiratory tube, or heat-moisture exchanger -HME) used on the ventilator and can be changed during normal ventilation or
short self test (SST). Changes in humidification type phase in at the start of
inspiration.
SST calibrates spirometry partly based on the humidification type. Changing
the humidification type without rerunning SST can affect the accuracy of spirometry and delivery.
The accuracy of the exhalation valve flow sensor varies depending on the water
vapor content of the expiratory gas, which depends on the type of humidification system in use. Because the temperature and humidity of gas entering the
expiratory filter differ based on the humidification type being used, spirometry
calculations also differ according to humidification type. For optimum accuracy, rerun SST to change the humidification type.
10.15.31 Humidifier

Volume

The dry, compressible volume in mL of the humidification chamber for the


humidification type entered during SST. Only applies if a humidifier is used.
10.16

Safety Net
While the ventilator is designed to be as safe and as reliable as possible, Covidien recognizes the potential for problems to arise during mechanical ventilation, either due to user error, patient-ventilator interactions, or because of
problems with the ventilator itself. Safety Net is a broad term that includes
strategies for handling problems that arise in the patient-ventilator system
(patient problems) as well as strategies to minimize the impact of system faults
on patient safety. In these scenarios, The ventilator is designed to alarm and to

Operators Manual

10-71

Theory of Operations

provide the highest level of ventilation support possible. in case of ventilator


malfunction. If the ventilator is not capable of ventilatory support, it opens the
patient circuit and allows the patient to breathe from room air if able to do so
(this emergency state is called Safety Valve Open (SVO). Safety mechanisms
are designed to be verified periodically or to have redundancy. The ventilator
is designed to ensure that a single-point failure does not cause a safety hazard
or affect its ability to annunciate a high-priority audible alarm.
10.16.1

User Error
The ventilator is designed to prevent the operator from implementing settings
that are clearly inappropriate for the patient's predicted body weight (PBW).
Each setting has either soft bounds (can be overridden) or hard bounds (no
override allowed) that alert the operator to the fact that the settings may be
inappropriate for the patient. In the event that the patient is connected
without any parameters being specified, the ventilator enters Safety PCV, a
safe mode of ventilation regardless of the circuit type in use (neonatal, pediatric, or adult) or patient's PBW. Safety PCV is entered after POST, if a patient
connection is made prior to settings confirmation. Safety PCV uses New Patient
default settings with exceptions shown in the following table:
Table10-9.Safety PCV Settings
Parameter
PBW

Neonatal: 3 kg
Pediatric:15 kg
Adult: 50 kg

mode

A/C

mandatory type

PC

fTOT (total respiratory rate)

Neonatal: 25 1/min
Pediatric: 16 1/min
Adult: 16 1/min

TI

Neonatal: 0.3 s
Pediatric: 0.7 s
Adult: 1 s

PI

15 cmH2O

O2%

10-72

Safety PCV value

Neonatal: 40%
Pediatric: 100%
Adult: 100%

Operators Manual

Short Self Test (SST)

Table10-9.Safety PCV Settings (Continued)


Parameter
PEEP
Trigger type

10.16.2

3 cmH2O
Neonatal: V-TRIG
Pediatric: P-TRIG
Adult: P-TRIG

PSENS

2 cmH2O

VSENS

1.0 L/min

1PPEAK

20 cmH2O

1VE TOT alarm

OFF

3VE TOT alarm

0.05 L/min

1VTE alarm

OFF

3VTE MAND alarm

OFF

3VTE SPONT alarm

OFF

Circuit type

Safety PCV value

Last set value, or adult if none available

Humidification type

Set value, or 'NON-HEATED EXP TUBE' if none


available

Humidifier volume

Last set value, or 480 mL if none available

Note:
In Safety PCV, expiratory pauses are not allowed.

Patient Related Problems


In case of patient problems, the ventilator remains fully operative and annunciates the appropriate alarm. The detection, response, and priority of each
patient-related alarm is determined by the actual patient problem. Reference
Alarms, p. 6-4 for a comprehensive description of the patient alarm system.

10.16.3

System Related Problems


The ventilator is designed to prevent system faults. Its modular design allows
the breath delivery unit (BDU) to operate independently of the graphical user

Operators Manual

10-73

Theory of Operations

interface (GUI) and several modules of the breath delivery sub-system have
redundancy that, if certain faults occur, provides for ventilatory support using
settings that do not depend on the suspect hardware. System faults include
the following:

10.16.4

Hardware faults (those that originate inside the ventilator and affect its performance)

Soft faults (faults momentarily introduced into the ventilator that interfere with
normal operation

Inadequate supply (AC power or external gas pressure)

Patient circuit integrity (occluded or disconnected circuit)

Background Diagnostic System


The ventilator has an extensive system of continuous testing processes. If an
error is detected in the background diagnostic system, the ventilator notifies
the operator by posting an entry in the diagnostic log. If the ventilator experiences an anomaly which causes an unintended reset, the ventilator will recover
from that reset and deliver a breath within three(3) s without any operator
intervention. After recovering from a reset, the ventilator uses the same settings that were in effect before the reset occurred.
The background test process compares monitored values of ventilator functions with expected values of ventilator sensors under normal conditions
regardless of whether the ventilator is in Stand-by or is ventilating a patient.
The ventilator will continue to ventilate the patient with the highest level of
support possible, and may revert to one of the states described. Reference
Ventilator Protection Strategies, p. 4-35.
Background tests include

Periodically initiated tests performed at intervals of a specific number of machine


cycles. These tests check hardware components directly affecting breath delivery,
safety mechanisms, and the GUI, and detect and correct corruption of control variable data.

Boundary checks performed at every analog measurement. These checks verify


measurement circuitry, including sensors.

Ventilation Assurance is a safety net feature invoked if the Background Diagnostics detect a problem with certain components in either the gas mix subsystem, the inspiratory subsystem, or the expiratory subsystem. Each subsystem
has a Backup Ventilation strategy that allows ventilation to continue by bypassing the suspect components giving the operator time to replace the ventilator.

10-74

Operators Manual

Short Self Test (SST)

Mix BUV is invoked if the measured gas mix is significantly different from the
set mix, if the accumulator pressure is out of range or if a fault is indicated in
the mix PSOLs or flow sensors. During MIX BUV, the normal mix controller is
bypassed and ventilation continues as set, except that the gas mix reverts to
100% Oxygen or Air, depending on where the fault indication was detected.
Backup circuits then control the pressure in the accumulator to keep it in the
proper range for the Inspiratory Module.
Inspiratory BUV is invoked if Background Diagnostics detect a problem in the
inspiratory module (PSOL and/or flow sensor signal out of range). In inspiratory
BUV, ventilation continues with the following settings:
Table10-10.Inspiratory Backup Ventilation Settings
Backup Ventilation parameter

Setting

PBW

Previously used setting during Vent Startup

Mode

A/C

Mandatory type

PC

Neonatal: 25 1/min
Pediatric: 16 1/min
Adult: 16 1/min

TI

Neonatal: 0.3 s
Pediatric: 0.7 s
Adult: 1 s

PI

15 cmH2O above PEEP

O2%

100% (21% if O2 not available)

PEEP

3 cmH2O

TPL

0s

Trigger type

V-TRIG; 2 L/min (adult/pediatric), 1.5 L/min (neonatal)

Gas flow

Controlled by pressure in the mix accumulator

During inspiratory BUV, the delivery PSOL is disabled, but gas delivery is
achieved via an inspiratory BUV solenoid valve, the gas flow being created by
pressure in the mix accumulator.
Exhalation BUV is invoked if problems with the Exhalation Valve driver are
detected. A backup analog circuit is enabled to control the exhalation valve

Operators Manual

10-75

Theory of Operations

though the more advanced control features (active exhalation valve control)
are not functional.

Note:
During Mix and Inspiratory BUV, gas supply to installed options is disabled.

Entry into BUV is logged in the alarm log and system diagnostic log, and the
status display provides an indicator that the ventilator is in BUV and which subsystem is affected.
When in BUV, a high priority alarm is annunciated, and the GUI displays an
alarm banner indicating BUV, blanks patient data, and a displays a pressure
waveform.
If the ventilator cannot provide any degree of reliable ventilatory support and
fault monitoring, then the ventilator alarms and enters the safety valve open
(SVO) emergency state. During SVO, the ventilator de-energizes the safety,
expiratory, and inspiratory valves, annunciates a high-priority alarm, and turns
on the SVO indicator. During SVO, a patient can spontaneously inspire room
air (if able to do so) and exhale. Check valves on the inspiratory and expiratory
sides minimize rebreathing of exhaled gas during SVO. During SVO the ventilator

Displays the elapsed time without ventilatory support

Does not display patient data (including waveforms)

Does not detect patient circuit occlusion or disconnect conditions

Visible indicators on the ventilator's GUI and status display illuminate when the
ventilator is in the SVO state. Other safeguards built into the ventilator include
a one-way valve (check valve) in the inspiratory pneumatic circuit allowing the
patient to inhale through the safety valve with limited resistance. This check
valve also limits exhaled flow from entering the inspiratory limb to reduce the
possibility of re-breathing exhaled CO2 gas.
10.17

Power On Self Test (POST)


Every time the ventilator is powered on or resets and at the beginning of Short
Self Test (SST) and Extended Self Test (EST) it performs Power On Self Test
(POST). POST checks the integrity of the GUI and Breath Delivery subsystems
and communication channels without operator intervention and takes approximately twelve(12) s to complete.

10-76

Operators Manual

Short Self Test (SST)

If POST detects a major fault, qualified service personnel must correct the
problem and successfully pass EST. Reference the Puritan Bennett 980 Series
Ventilator Service Manual for more details on POST.
10.18

Short Self Test (SST)

10.19

SST is a short (about 5 minutes) and simple sequence of tests that verifies
proper operation of breath delivery hardware (including pressure and flow
sensors), checks the patient circuit (including tubing, humidification device,
and filters) for leaks, and measures the circuit compliance and resistance.
SST also checks the resistance of the expiratory filter. SST, in normal mode,
can only be performed at start up, prior to initiation of ventilation. Covidien
recommends running SST every 15 days, between patients, and when
changing the patient circuit or its configuration (including changing circuit
type, adding or removing in-line water traps, or using a different type or
style of patient circuit). Reference To run SST, p. 3-48. The ventilator does not
allow access to SST if it senses a patient is connected.

Extended Self Test (EST)


EST verifies the integrity of the ventilators subsystems using operator participation. EST requires a gold standard test circuit and a stopper to block the
patient wye. All test resources, including the software code to run EST, exist in
the ventilator. EST testing, excluding tests of optional equipment (such as the
extended battery) takes about ten (10) minutes. Reference EST (Extended Self
Test), p. 3-52.

WARNING:
Do not enter Service mode with a patient attached to the ventilator. Serious
injury could result.

Operators Manual

10-77

Theory of Operations

Page Left Intentionally Blank

10-78

Operators Manual

11 Specifications
11.1Overview
This chapter contains the following specifications for the Puritan Bennett
980 Series Ventilator:

Physical

Electrical

Interface

Environmental

Performance (Ranges, resolution, and accuracies for ventilator settings, alarm settings, and patient data)

Regulatory Compliance

WARNING:
Due to excessive restriction of the Air Liquide, SIS, and Drger hose
assemblies, reduced ventilator performance levels may result when oxygen
or air supply pressures < 50 psi (345 kPa) are employed.

11.2Measurement Uncertainty
Table11-1.Performance Verification Equipment Uncertainty
Measured Parameter

Offset

Gain

Flow

0.1001 SLPM

2.7642 % reading

Pressure

0.121594 cmH2O

0.195756 % reading

Oxygen Concentration

0.0168 % O2

0.0973 % reading

Temperature

0.886041 C

0.128726 % reading

Atmospheric Pressure

1.76 cmH2O

Specifications

During breath delivery performance verification for flow and pressure based
measurements, the equipment inaccuracy is subtracted from the acceptance
specification as follows:

Net Acceptance Gain = Requirement Specification Gain - Measurement Uncertainty Gain

Net Acceptance Offset = Requirement Specification Offset - Measurement Uncertainty Offset

Acceptance Limit = [(Net Acceptance Offset) + (Net Acceptance Gain) x (Setting)]

(Setting- Acceptance Limit) Measurement (Setting + Acceptance Limit)

For derived parameters, such as volume, compliance, etc., the individual sensor
uncertainties are combined and applied as applicable to determine the acceptance limits.

11.3Physical Characteristics
Table11-2.Physical Characteristics

Weight

113 lb (51.26 kg) including BDU, GUI, standard base, and primary
battery
BDU only: 69 lb (31.3 kg)

Dimensions

Ventilator: 12.5 width x 11.5 depth x 43.5 height


(32 cm x 30 cm x 111 cm) (not including GUI screen)
Ventilator: 12.5 width x 11.5 depth x 58 height
(32 cm x30 cm x 148 cm) (including GUI screen
Standard base: 22.5 width x 26 depth (58 cm x 66 cm)

A-weighted sound pressure


level, ventilator (average)

At a distance of one (1) meter, does not exceed 48 dBA at 5 L/min

A-weighted sound power level,


ventilator

Does not exceed 58 dBA below 500 mL/min

Connectors

Inspiratory and expiratory limb connectors are 22 mm OD conical


fittings compliant with ISO 5356-1

Inspiratory/ Expiratory filters

Reference filter Instructions For Use for complete specifications

Pressure units (chosen by operator)

Hectopascal (hPa)
centimeters of water (cmH2O)

Displayed weight units

Kilograms (kg) or Pounds (lb) (user selectable)

Displayed length units

Centimeters (cm) or Inches (in) (user selectable)

Operator's Manual

Physical Characteristics

Table11-3.Pneumatic Specifications
Oxygen and air inlet supplies

Pressure: 241 to 600 kPa (35 psi to 87 psi)


Flow: Maximum of 200 L/min

Oxygen sensor life

Up to one year. Operating life varies depending on oxygen usage


and ambient temperature.

Gas mixing system

Range of flow from the mixing system:


Up to 150 L/min for Adult patients. Additional flow is available
(peak flow to 200 L/min) for compliance compensation
Up to 80 L/min for pediatric circuit type
Up to 30 L/min for neonatal circuit type
Leakage from one gas system to another: Meets IEC 80601-2-12
standard
Operating pressure range: 35 psi to 87 psi (241 to 600 kPa)

Table11-4.Technical Specifications
Maximum limited pressure (PLIMmax)

A fixed pressure limit to the safety valve limits


circuit pressure to < 125 cmH2O (123 hPa) at the
patient wye.

Maximum working pressure (PWmax)

PWmax is ensured by the high pressure limit


(2PPEAK) when PI is 125 cmH2O (98.07 hPA)

Response time to change in FiO2 setting from


21% to 90% O2 (measured at the patient wye)

< 18 s for volumes > 150 mL


< 19 s for volumes 30 mL but 150 mL
< 20 s for volumes 2 mL but < 30 mL

Measuring devices

Pressure measurements:
Type: Solid stated differential pressure transducer
Sensing position: Inspiratory module; expiratory
module
Mean circuit pressure (PMEAN): -20 cmH2O (-20
hPa) to 130 cmH2O (127 hPa)
Peak circuit pressure (PPEAK): -20 cmH2O (-20
hPa) to 130 cmH2O (127 hPa)
Volume Measurements:
Type: Hot film anemometer
Sensing position: Inspiratory module; expiratory
module
Oxygen measurement:
Type: Galvanic cell
Sensing position: Inspiratory module

Minute volume (VE TOT) capability, ventilator

Up to 75 L/min

Results of ventilator testing using circuits identified for use with the ventilator system

Operator's Manual

Specifications

Table11-4.Technical Specifications (Continued)

Internal Inspiratory filter bacterial/viral filtration


efficiency

> 99.999%

Internal Inspiratory filter particle filtration efficiency

> 99.97% retention of particles 0.3 m nominal


at 100 L/min flow

Internal Inspiratory filter resistance

0.2 cmH2O < resistance < 2.2 cmH2O at 30 L/min


flow
0.2 cmH2O < resistance < 1.7 cmH2O at 15 L/min
flow

External Inspiratory filter resistance

0.2 cmH2O < resistance < 2.2 cmH2O at 30 L/min


flow
0.2 cmH2O < resistance < 2.2 cmH2O at 15 L/min
flow

Combined inspiratory filter resistance

0.2 cmH2O < resistance < 5.5 cmH2O at 30 L/min


flow
0.2 cmH2O < resistance < 1.7 cmH2O at 15 L/min
flow

External Inspiratory filter resistance

0.2 cmH2O < resistance < 2.2 cmH2O at 30 L/min


flow
0.2 cmH2O < resistance < 1.7 cmH2O at 15
L/ min flow

External Inspiratory filter bacterial/viral filtration


efficiency, reusable

> 99.999%

External Inspiratory filter particle filtration efficiency, reusable

> 99.97% retention of particles 0.3 m nominal


at 100 L/min flow

External Inspiratory filter resistance, reusable

0.2 cmH2O < resistance < 4.2 cmH2O at 60 L/min


0.2 cmH2O < resistance < 2.2 cmH2O at 30 L/min
0.2 cmH2O < resistance < 1.7 cmH2O at 15 L/min

External Inspiratory filter bacterial/viral filtration


efficiency, disposable

> 99.999%

External Inspiratory filter particle filtration efficiency, disposable inspiratory filter

> 99.97% retention of particles 0.3 m nominal


at 100 L/min flow

Expiratory filter particle filtration efficiency, reusable

> 99.97% retention of particles 0.3 m nominal


at 100 L/min flow

Expiratory filter bacterial/viral filtration efficiency,


reusable

> 99.999%

Expiratory filter resistance (pediatric/adult, reusable and disposable)

< 2.5 cmH2O at 30 L/min when new


< 1.7 cmH2O at 15 L/min

Operator's Manual

Physical Characteristics

Table11-4.Technical Specifications (Continued)


Expiratory filter bacterial/viral filtration efficiency,
disposable

> 99.999%

Expiratory filter particle filtration efficiency, disposable

> 99.97% retention of particles 0.3 m nominal


at 100 L/min flow

Expiratory filter bacterial/viral filtration efficiency


neonatal, disposable)

> 99.999%

Expiratory filter particle filtration efficiency (neonatal, disposable)

> 99.97% retention of particles 0.3 m nominal


at 100 L/min flow

Expiratory filter resistance (neonatal, disposable)

< 0.58 cmH2O at 2.5 L/min

Circuit compliance

NEONATAL: < 0.25 mL/cmH2O to > 3.0 mL/


cmH2O
PEDIATRIC: 1.05 mL/cmH2O to 9 mL/cmH2O
ADULT: 1.05 mL/cmH2O to 12 mL/cmH2O

Inspiratory limb circuit resistance

NEONATAL: 0.2 cmH2O/L/s to 3.5 cmH2O/L/s


PEDIATRIC: 0.2cmH2O to 7.5 cmH2O/L/s
ADULT: 0.2cmH2O to 12.5 cmH2O/L/s

Expiratory limb circuit resistance

NEONATAL: 0.2 cmH2O/L/s to 3.5 cmH2O/L/s


PEDIATRIC: 0.2cmH2O to 7.5 cmH2O
ADULT: 0.2cmH2O to 12.5 cmH2O

Alarm volume (primary)


Measurement uncertainty: 3 dBA

High priority alarm volume range (dBA): 58


(volume setting 1) to 86 (volume setting 10)
Medium priority alarm volume range (dBA): 52
(volume setting 1) to 78 (volume setting 10)
Low priority alarm volume range (dBA): 50
(volume setting 1) to 76 (volume setting 10)
Measured 1 m from front, rear, and sides of ventilator
Reference Alarm Volume Key, p. 6-8 for alarm
volume behavior during an alarm condition.

Alarm volume (secondary)


Measurement uncertainty: 3 dBA

Minimum 64 dBA measured 1 m from front, rear,


and sides of ventilator.

Operator's Manual

Specifications

11.4Electrical Specifications
Table11-5.Electrical Specifications
Electrical ratings

100V ~, 50-60 Hz, 2.25A


120 V ~, 50-60 Hz 1.5A
220-240 V ~, 50-60 Hz, 0.75A

Mains overcurrent release

CB1: 4A

Earth leakage current

300 A

Touch current

100 A

Patient Leakage current

100 A maximum

11.5Interface Requirements
Table11-6.Interface Pin Designations
Pin

Signal

Name

N/C

Not connected

RxD

Receive data

TxD

Transmit data

N/C

Not connected

GND

Ground

N/C

Not connected

RTS

Request to send

CTS

Clear to send

N/C

Mot connected

The pin-out for the nurse call interface is as follows:

Operator's Manual

Environmental Specifications

Table11-7.Nurse Call Pin Designations


Pin

Configuration

Normally closed (NC)

Relay common

Normally open (NO)

Not connected

11.6Environmental Specifications

Table11-8.Environmental Specifications
Specification

Operation

Storage

Temperature

10C to 40C (50F to 104F) Ventilator


10C to 35C (50F to 95F) Internal
Battery Charger

-20C to 70C (-68F to 158F)

70kPa to 106 kPa (10.15 psi to (15.37


psi)

50 kPa to 106 kPa (7.25 psi to


15.37 psi)

-411.5 m to 3048 m (-1350 ft to 10000


ft)

6096 m max (20000 ft max)

10% to 95% non-condensing

10 to 95% non-condensing

Atmospheric Pressure
Altitude
Relative Humidity

Note:
The limits marked on the device label represent out-of-box storage conditions as
follows:

Temperature: (10C to 40C (50F to 104F)

Pressure: 70kPa to 106 kPa (10.15 psi to 15.37 psi)

Relative Humidity: 10% to 95% non-condensing

Operator's Manual

Specifications

11.7Performance Specifications
11.7.1Ranges and Resolutions
Reference the table below for ranges, resolutions, accuracies, new patient
default values, and dependencies, where applicable, for ventilator settings.
Reference Alarm Settings Range and Resolution, p. 11-17 for alarm settings,
and Reference Patient Data Range and Resolution, p. 11-19 for displayed
patient data parameters.
Table11-9.Ventilator Settings Range and Resolution
Setting

Description

Range and Resolution

Apnea ventilation

A safety mode of ventilation that starts if the


patient does not receive a
breath for an elapsed time
exceeding the apnea interval.

See individual apnea settings.

Apnea expiratory time (TE)

For mandatory PC apnea


breaths, the time interval
between the end of inspiration and the beginning
of the next inspiration.

Range:0.20 s to 59.8 s
Resolution: 0.01 s

Apnea I:E ratio

In PC breath types, specifies the ratio of apnea


inspiratory time to apenea
expiratory time.

Range: I:E 1.00:1


Resolution: 0.01 for values >
1:10.0; 0.1 for values 1:10 and >
1:100; 1 for values 1:100

Apnea flow pattern

The flow shape of the


delivered mandatory
volume-based (VC) apnea
breath.

Range: SQUARE, descending ramp

Apnea inspiratory pressure (PI)

The pressure above PEEP


at which gas is delivered to
the patient during mandatory PC apnea breaths.

Range: 5 cmH2O to 90-PEEP cmH2O


Resolution: 1 cmH2O

Apnea inspiratory time (TI)

Same as inspiratory time


for non-apnea ventilation

Range: 0.20 s to 8 s
Resolution: 0.01 s in PC or VC+,
0.02 s in VC

Operator's Manual

Performance Specifications

Table11-9.Ventilator Settings Range and Resolution (Continued)


Setting

Operator's Manual

Description

Range and Resolution

Apnea interval (TA)

The time after which the


ventilator transitions to
apnea ventilation
TA 60/fA

Range: 10 s to 60 s or OFF in CPAP


Resolution: 1 s

Apnea O2 %

Determines the oxygen


concentration in a standard mixture of air and
oxygen

Range: 21%O2 to 100% O2


Resolution: 1%

Apnea peak inspiratory


flow (VMAX)

The maximum rate of tidal


volume delivery during
mandatory volume-based
apnea breaths.

Range: When mandatory type is VC:


NEONATAL: 1 L/min to 30 L/min
PEDIATRIC: 3.0 L/min to 60 L/min
ADULT: 3.0 L/min to 150 L/min
Resolution: 0.1 L/min for flows < 20
L/min (BTPS); 1 L/min for flows 20
L/min (BTPS)

Apnea respiratory rate (fA)

Sets the number of


volume- or pressure-based
breaths per minute for
ventilator initiated mandatory (VIM) apnea breaths

Range: 2.0 1/min to 40 1/min


Resolution: 0.1 1/min for 2.0 to 9.9;
1 1/min for 10 1/min to 40 1/min

Apnea tidal volume (VT)

Sets the volume of gas


delivered to the patients
lungs during a mandatory,
volume-controlled apnea
breath. Apnea tidal
volume is compensated
for body temperature and
pressure, saturated (BTPS)
and the compliance of the
patient circuit.

Range:
NEONATAL: 3 mL to 315 mL
PEDIATRIC/ADULT: 25 mL to
2500 mL
Resolution: 0.1 mL < 5 mL; 1 mL for
5 mL to 100 mL; 5 mL for100 mL to
395 mL; 10 mL for values 400 mL

Apnea constant during


rate change

Specifies which of the


three operator-adjustable
breath timing variables
remains constant when
respiratory rate is changed
during apnea ventilation.

Range: TI

Apnea mandatory type

The type of mandatory


breath delivered during
apnea ventilation

Range: PC, VC

Circuit type

Specifies the circuit for


which compliance and
resistance values during
SST have been calculated

Range: NEONATAL, PEDIATRIC,


ADULT

Specifications

Table11-9.Ventilator Settings Range and Resolution (Continued)


Setting

10

Description

Range and Resolution

Constant during rate


change

Specifies which of the


three operator-adjustable
breath timing variables
remains constant when
respiratory rate is
changed.

Range: I:E ratio, TI, TE for PC or VC+


breaths; TH:TL ratio, TH,TL in BiLevel

Disconnect sensitivity
(DSENS)

The percentage of
returned volume lost,
above which the ventilator
declares a circuit disconnect alarm when Leak
Sync is not enabled or
installed.

Range: (if Leak Sync not enabled)


20% to 95% or OFF
Range: (if Leak Sync enabled)
NEONATAL: 1 L/min to 15 L/min
PEDIATRIC: 1 L/min to 40 L/min
ADULT: 1L/min to 65 L/min
Resolution: (Leak Sync not enabled)
1%
Resolution: (Leak Sync enabled) 0.5
L/min for values< 10 L/min; 1 L/min
for values 10 L/min

Expiratory sensitivity
(ESENS)

The percentage of VMAX


that, when reached,
causes the ventilator to
cycle from inspiration to
exhalation during spontaneous, pressure-based
breaths

Range: 1% to 80%
1 L/min to 10 L/min when Spontaneous Type is PAV+.

Expiratory time (TE)

For PC or VC+ breaths, the


time interval between the
end of inspiration and the
beginning of the next
inspiration. The end of the
exhalation phase is considered to be when the flow
rate at the patient wye
remains less than 0.5
L/min above the base flow.

Range: 0.20 s
Resolution: 0.01 s

Flow pattern

The flow shape of the


delivered mandatory or VC
breath

Range: SQUARE, descending ramp

NOTE: Default value is not expected


to need adjustment. Only adjust after
becoming experienced with PAV+
and only if it is suspected that the
ventilator is not cycling at the
patients end-of-inspiration.

Operator's Manual

Performance Specifications

Table11-9.Ventilator Settings Range and Resolution (Continued)


Setting

Operator's Manual

Description

Range and Resolution

Flow sensitivity (VSENS)

For flow triggered breaths,


determines the volume of
flow (below the base flow)
required to begin a mandatory or spontaneous
patient initiated breath.

Range:
NEONATAL: 0.1 L/min to 10 L/min
PEDIATRIC/ADULT: 0.2 L/min to
20.0 L/min
Resolution: 0.1 L/min

Gender

The patients gender

Range: Male or Female

Height

The patients height

Range: 19.5 cm to 280 cm; 7.5 in to


110 in
Resolution: 0.5 cm for heights < 35
cm; 1 cm for heights < 254 cm; 2 cm
for heights 254 cm; 0.25 in for
heights < 14 in; 0.5 in for heights
<100 in; 1 in for heights 100 in
Reference Predicted Body Weight
(PBW) Calculation, p. 4-21.

High spontaneous inspiratory time limit (2TI SPONT)

Active in NIV only, allows


the operator to select the
maximum spontaneous
inspiratory time.

Range:
NEONATAL: 0.2 s to 1.7 s
PEDIATRIC/ADULT: 0.4 s to 5 s

Humidification type

The type of humidification


system used on the ventilator

Range: HME, non-heated expiratory


tube, heated expiratory tube

Humidifier volume

The empty fluid volume of


the currently installed
humidifier.

Range: 100 mL to 1000 mL


Resolution: 10 mL

Elevate O2%

The percentage of O2 to
be added to the current
air/O2 mixture for two
minutes

Range: 1% to 100%
Resolution: 1% between 1% and
10; 5% between 5% and 75%;
jumps to 100% when increased
above 75%

I:E ratio

In PC and VC+ breath


types, specifies the ratio of
inspiratory time to expiratory time.

Range: 1:299 to 149:1


Resolution: 0.01 for values > 1:10;
0.1 for values 1:10.0 and >
1:100.0; 1 for values 1:100
Displayed as XX:1 when I:E 1; displayed as 1:XX when I:E < 1

Inspiratory pressure (PI)

The pressure above PEEP


at which gas is delivered to
the patient during mandatory PC breaths.

Range: 5 cmH2O to 90 cmH2O


Resolution:1 cmH2O

11

Specifications

Table11-9.Ventilator Settings Range and Resolution (Continued)


Setting

12

Description

Range and Resolution

Inspiratory time (TI)

The time during which an


inspiration is delivered to
the patient during mandatory PC or VC+ breaths.

Range: 0.2 s to 8 s for mandatory


PC, and VC+ breaths, (TPL+ 0.2 s to 8
s in VC)
Resolution: 0.01 s for PC or VC+
breaths; 0.02 s for VC breaths

Leak Sync (leak compensation)

Compensates for leaks


during INVASIVE or noninvasive (NIV) ventilation.

Range: Enabled or Disabled

Mandatory type

The type of mandatory


breath delivered in A/C,
SPONT or SIMV modes.
SPONT mode allows mandatory type selection for
operator initiated mandatory (OIM) breaths.

Range: PC, VC, VC+

mL/kg ratio

The default tidal volume/


PBW ratio (only adjustable
in Service Mode)

Range: 5.0 mL/kg to 10 mL/kg


Resolution: 0.5 mL/kg

Operator's Manual

Performance Specifications

Table11-9.Ventilator Settings Range and Resolution (Continued)


Setting

Operator's Manual

Description

Range and Resolution

Mode

The ventilation mode. The


mode determines the
allowable breath types:
A/C assist/control a
mandatory mode allowing
volume controlled (VC),
pressure controlled (PC),
or VC+ breath types.
SPONT allows the
patient to initiate the
breath. Applicable SPONT
breath types are pressure
support (PS), volume
support (VS), tube compensated (TC) or PAV+ if
the PAV+ option is
installed.
SIMV Synchronized
Intermittent Mandatory
Ventilation a mixed ventilatory mode providing
mandatory breaths and
allowing a patient spontaneous breaths during the
breath cycle.
BiLevel a mixed ventilatory mode combining the
attributes of both mandatory and spontaneous
breaths incorporating two
pressure levels, PH and PL

Range: A/C, SPONT, SIMV, BiLevel


(if option installed but not available
when vent type is NIV; CPAP (only
available when circuit type is NEONATAL and vent type is NIV))

O2% (delivered)

Percentage of delivered
oxygen in the gas mixture

Range: 21% to 100%


Resolution: 1%

Peak inspiratory flow


(VMAX)

The maximum rate of tidal


volume delivery during
mandatory volume-based
breaths.

Range: When mandatory type is VC:


NEONATAL: 1 L/min to 30 L/min
PEDIATRIC: 3.0 L/min to 60 L/min
ADULT: 3.0 L/min to 150 L/min
Resolution: 0.1 L/min for values <
20 L/min (BTPS); 1 L/min for values
20 L/min (BTPS)

PEEP

Sets the positive end-expiratory pressure, defined as


the pressure targeted in
the patient circuit during
exhalation.

Range: 0 cmH2O to 45 cmH2O


Resolution: 0.5 cmH2O from 0.0 to
19.5 cmH2O; 1 cmH2O from 20 to
45 cmH2O

13

Specifications

Table11-9.Ventilator Settings Range and Resolution (Continued)


Setting

14

Description

Range and Resolution

PH

The positive pressure


during the insufflation
phase in BiLevel ventilation.

Range: 5 cmH2O to 90 cmH2O


Resolution: 1 cmH2O

PL

The positive pressure in


the patient circuit during
the expiratory phase of
BiLevel ventilation.

Range: 0 cmH2O to 45 cmH2O


Resolution: 0.5 cmH2O from 0.0 to
19.5 cmH2O; 1 cmH2O from 20 to
45 cmH2O

Plateau time (TPL)

The amount of time inspiration is held in the


patients lungs after inspiratory flow ceases for
volume-based mandatory
breaths. Considered part
of inspiratory phase for I:E
ratio calculations.

Range: 0s to 2 s
Resolution: 0.1 s

Predicted Body Weight


(PBW)

Indicates an approximation of the patients


ideal body weight
based upon their gender
and height (or length for
neonatal patients). PBW
determines default limits
and limits for breath delivery parameters.

Range:
NEONATAL: 0.3kg (0.66 lb) to 7.0
kg (15 lb) when NeoMode 2.0 option
is installed;
PEDIATRIC: 3.5 kg (7.7 lb) to 35 kg
(77 lb)
ADULT: 25 kg (55.12 lb)
Resolution: 0.01 kg for weights < 1
kg, 0.1 kg for weights 1 kg and <
10 kg, 1 kg for weights 10 kg

Pressure sensitivity (PSENS)

For pressure triggered


breaths, determines the
amount of pressure below
PEEP required to begin a
mandatory or spontaneous patient initiated
breath.

Range: 0.1 cmH2O to 20.0 cmH2O


Resolution: 0.1 cmH2O

Pressure support (PSUPP)


or PS

The positive pressure


above PEEP (or PL in BiLevel) during a spontaneous
breath.

Range: 0 cmH2O to 70cmH2O


Resolution: 1 cmH2O

Respiratory rate (f)

Sets the number of


volume- or pressure-based
breaths per minute for
ventilator initiated mandatory (VIM) breaths in A/C,
SIMV, and BiLevel modes

Range:
NEONATAL: 1.0 1/min to 150 1/min
PEDIATRIC/ADULT:
Resolution: 0.1 from 1.0 1/min to
9.9 1/min; 1 1/min from 10 1/min to
150 1/min

Operator's Manual

Performance Specifications

Table11-9.Ventilator Settings Range and Resolution (Continued)


Setting

Operator's Manual

Description

Range and Resolution

Rise time %

Sets the speed at which


inspiratory gas delivered to
the patient reaches the
pressure target in BiLevel,
PC, VC+, VS, or PS. Higher
percentages of rise time
produce inspiratory pressure trajectories with
shorter time to the target
value.

Range: 1% to 100%
Resolution: 1%

Spontaneous type

The breath type for patient


initiated spontaneous
breaths in SIMV, SPONT,
and BiLevel modes.

Range: PS, TC, PAV+, or VS

% Supp

In Tube Compensation,
specifies the additional
positive pressure desired
to overcome resistance of
the artificial airway.

Range:10% to 100%
Resolution: 5%

% Supp

In PAV+, specifies the percentage of total inspiratory work of breathing


(WOB) performed by the
ventilator.

Range: 5% to 95%
Resolution: 5%

TH (time high)

The duration of the insufflation phase during


BiLevel ventilation.

Range: 0.2 s to 30 s
Resolution: 0.01 s

TL (time low)

The duration of the expiratory phase during BiLevel


ventilation.

Range: 0.20 s
Resolution: 0.01 s

TH:TL ratio

In BiLevel, specifies the


ratio of insufflation time to
expiratory time

Range: 1:299 to 4:1; in BiLevel TH:TL


Resolution: 0.01 for < 10.00:1 and
> 1:10.00; 0.1for [< 100.0:1 and
10.0:1] or[ 1:10.0 and > 1:100.0]; 1
for < 1:100.0 or 100:1

15

Specifications

Table11-9.Ventilator Settings Range and Resolution (Continued)


Setting

16

Description

Range and Resolution

Tidal volume (VT)

The volume of gas delivered to the patient during


a mandatory volumebased breath. VT compensates for body temperature and pressure,
saturated (BTPS) and
circuit compliance. Applicable for volume-based
breaths.

Range:
NEONATAL: 2 mL to 315 mL
PEDIATRIC: 25 mL to 1590 mL
ADULT: 25 mL to 2500 mL
Resolution: 0.1 mL < 5 mL; 1 mL for
5 mL and < 100 mL; 5 mL for 100
mL to 395 mL; 10 mL for values
400 mL

Volume support (VT SUPP)


or VS

The volume of gas delivered to the patient during


spontaneous, volume supported breaths

Range:
NEONATAL: 2 mL to 310 mL
PEDIATRIC: 25 mL to 1590 mL
ADULT: 25 mL to 2500 mL
Resolution: 0.1 mL for 5 mL; 1 mL
for 5 mL to < 100 mL; 5 mL for 100
mL to < 400 mL; 10 mL 400 mL

Trigger type

Determines whether flow


changes (V-TRIG), pressure
changes (P-TRIG), or
intrapleural pressure
changes (IE Sync) trigger
patient breaths

Range:
NEONATAL: V-TRIG
PEDIATRIC/ADULT: V-TRIG, P-TRIG or
IE Sync in SPONT (if IE Sync option
installed)

Tube ID

The internal diameter of


the artificial airway used to
ventilate the patient.

Range: 4.5 mm to 10 mm when


spontaneous type is TC
Range: 6 mm to 10 mm when spontaneous type is PAV+
Resolution: 0.5 mm

Tube type

The type of artificial airway


used to ventilate the
patient.

Range: Endotracheal (ET), tracheal


(Trach)

Ventilation type

Invasive or non-invasive
(NIV) ventilation type
based upon the type of
breathing interface used.
Invasive: ET or Trach tubes
NIV: masks, infant nasal
prongs, or uncuffed ET
tubes

Range: INVASIVE, NIV

Operator's Manual

Performance Specifications

Table11-10.Alarm Settings Range and Resolution


Setting

Range and Resolution

Alarm volume

Controls the volume of alarm


annunciations

Range: 1 (minimum) to 10
(maximum)
Resolution: 1

Apnea Interval (TA)

The Apnea alarm condition


indicates that neither the
ventilator nor the patient has
triggered a breath for the
operator-selected Apnea
Interval (TA). When the
Apnea alarm condition is
true, the ventilator invokes
mandatory ventilation as
specified by the operator.

Range: 10 s to 60 s or OFF in
CPAP
Resolution: 1 s

High circuit pressure setting


(2PPEAK)

The 2PPEAK alarm indicates


the patients airway pressure
the set alarm level

Range: 7 cmH2O to 100


cmH2O
Resolution: 1 cmH2O

Low circuit pressure setting

The 3PPEAK alarm indicates


the measured airway pressure the set alarm limit
during an NIV or VC+ inspiration.

Range:
NIV: OFF or 0.5 cmH2O to
<100 cmH2O
Resolution: 0.5 cmH2O for
values < 20.0 cmH2O;
1 cmH2O for values 20
cmH2O

The 1VE TOT alarm indicates


the measured total minute
volume the set alarm limit.

Range: OFF and


NEONATAL: 0.1 L/min to 10
L/ min
PEDIATRIC: 0.1 L/min to 30
L/min
ADULT: 0.1 L/min to 100
L/min
Resolution: 0.005 L/min for
values < 0.50 L/min; 0.05
L/min for values 0.5 L/min to
< 5.0 L/min; 0.5 L/min for
values 5.0 L/min

(4PPEAK)

High exhaled minute volume


alarm setting (2VE TOT)

Operator's Manual

Description

17

Specifications

Table11-10.Alarm Settings Range and Resolution (Continued)


Setting

18

Description

Range and Resolution

High exhaled tidal volume


alarm setting (2VTE)

The 1VTEalarm indicates that


the measured exhaled tidal
volume the set alarm limit
for spontaneous and mandatory breaths.

Range: OFF and


NEONATAL: 5 mL to 500 mL
PEDIATRIC: 25 mL to 1500
mL
ADULT: 25 mL to 3000 mL
Resolution: 1 mL for values <
100 mL; 5 mL for values 100
mL and < 400 mL; 10 mL for
values 400 mL

High inspired tidal volume


alarm limit (2VTI)

The 1VTI alarm indicates the


delivered volume of any
breath the set alarm limit.

Range: 6 mL to 6000 mL
Resolution: 1 mL for values <
100 mL; 5 mL for values 100
mL to < 400 mL; 10 mL for
values 400 mL

High respiratory rate alarm


setting (2fTOT)

The 1fTOT alarm indicates the


measured breath rate the
set alarm limit.

Range: OFF or
NEONATAL: 10 1/min to 170
1/min
PEDIATRIC/ADULT: 10 1/min
to 110 1/min
Resolution: 1 1/min

High spontaneous inspiratory


time limit (2TI SPONT)

The 2TI SPONT indicator allows


the operator to select the
maximum spontaneous inspiratory time of an NIV breath.
No alarm is annunciated; only
the symbol 2TI SPONT appears
on the screen near the NIV
indicator when inspiration
time exceeds the setting. If
2TI SPONT is exceeded, the
ventilator transitions from
inspiration to exhalation.

Range:
NEONATAL: 0.2 s to the
value of the NIV inspiratory
time limit trigger for the
patients PBW and circuit type
s
PEDIATRIC/ADULT: 0.4 s to
the value of the NIV inspiratory
time limit trigger for the
patients PBW and circuit type
s
Resolution: 0.1 s

Low exhaled mandatory tidal


volume alarm setting (4VTE
MAND)

The 3VTE MAND alarm indicates the measured mandatory tidal volume the set
alarm limit.

Range: OFF and


NEONATAL: 1 mL to 300 mL
PEDIATRIC: 1 mL to 1000 mL
ADULT: 1 mL to 2500 mL
Resolution: 1.0 mL for values
< 100 mL; 5 mL for values
100 mL and < 400 mL; 10 mL
for values 400 mL

Operator's Manual

Performance Specifications

Table11-10.Alarm Settings Range and Resolution (Continued)


Setting

Description

Range and Resolution

Low exhaled minute volume


alarm setting (4VE TOT)

The 4VE TOT alarm indicates


the measured exhaled minute
volume the set alarm limit
for mandatory and spontaneous breaths.

Range: OFF when vent type =


NIV and
NEONATAL: 0.01 L/min to 10
L/min
PEDIATRIC: 0.05 L/min to
30 L/min
ADULT: 0.05 L/min to
60 L/min
Resolution: 0.005 L/min for
values < 0.50 L/min;
0.05 L/min for values 0.50
L/min and < 5.0 L/min; 0.5
L/min for values > 5.0 L/min

Low exhaled spontaneous


tidal volume alarm setting
(4VTE SPONT)

The 3VTE SPONT alarm indicates the measured spontaneous tidal volume the set
alarm limit.

Range: OFF and


NEONATAL: 1 mL to 300 mL
PEDIATRIC: 1 to 1000 mL
ADULT: 1 to 2500 mL
Resolution: 1 mL for values <
100 mL; 5 mL from 100 mL to
< 400 mL; 10 mL for values
400 mL

Table11-11.Patient Data Range and Resolution


Data Value

Operator's Manual

Description

Range and Resolution

Breath phase

The breath phase indicator


displays the breath delivery
phase (inspiration or exhalation) currently being delivered
to the patient.

Range: Control (C), Assist (A),


Spontaneous (S)

Inspired tidal volume (VTL)


during Leak Sync

The volume inspired for each


breath when Leak Sync is
enabled.

Range: 0 mL to 6000 mL
Resolution: 0.1 mL for values
< 10 mL; 1 mL for values 10 mL
to 6000 mL

Inspired tidal volume (VTI)

The volume inspired for a


pressure- based breath

Range: 0 mL to 6000 mL
Resolution: 0.01 mL for 0 mL
to 9.9 mL, 1 mL for values 10
mL to 6000 mL

Inspired tidal volume


(VTL)during Leak Sync

The volume inspired for each


breath when Leak Sync is
enabled.

Range: 0 mL to 6000 mL
Resolution: 1 mL for values <
10 mL; 1 mL for values 10 mL
to 6000 mL

19

Specifications

Table11-11.Patient Data Range and Resolution (Continued)


Data Value

20

Description

Range and Resolution

Dynamic compliance (CDYN)

The result of dividing the


delivered tidal volume by the
peak airway pressure.

Range: 0 mL/cmH2O to 200


mL/ cmH2O
Resolution: 0.1 mL/cmH2O for
values < 10 mL/cmH2O; 1 mL/
cmH2O for values 10 mL/
cmH2O

Dynamic resistance (RDYN)

The change in pressure per


unit change in flow.

Range: 0.0 cmH2O/L/s to 100


cmH2O/L/s
Resolution: 0.1 cmH2O/L/ for
values < 10 cmH2O/L/s; 1
cmH2O/ L/s for values 10
cmH2O/L/s

End expiratory flow (EEF)

The rate of expiratory flow


occurring at the end of exhalation.

Range: 0 to 150 L/min


Resolution: 0.1 L/min for
values < 20 L/min; 1 L/min for
values 20 L/min

End expiratory pressure (PEEP)

The pressure at the end of the


expiratory phase of the previous breath (also applies in
BiLevel).

Range: -20.0 cmH2O to 130


cmH2O
Resolution: 0.5 cmH2O
between -10.0 and +10.0
cmH2O; 1 cmH2O for values 10 cmH2O and 10 cmH2O

End inspiratory pressure


(PI END)

The pressure at the end of the


inspiratory phase of the
current breath (also applies in
BiLevel).

Range: -20.0 cmH2O to 130


cmH2O
Resolution: 0.1 cmH2O for 20.0 cmH2O to 9.9 cmH2O; 1.0
cmH2O for values 10 cmH2O to
130 cmH2O

Exhaled mandatory tidal


volume (VTE MAND)

The exhaled volume of the


last mandatory breath. When
the mode is SPONT, and no
mandatory breaths have
occurred for a time period 2
minutes, the VTE MAND indicator is hidden. Mandatory
breaths can occur during
SPONT mode via manual
inspiration.

Range: 0 mL to 6000 mL
Resolution: 0.1 mL for 0 mL to
9.9 mL; 1 mL for 10 mL to 6000
mL

Operator's Manual

Performance Specifications

Table11-11.Patient Data Range and Resolution (Continued)


Data Value

Operator's Manual

Description

Range and Resolution

Exhaled minute volume


(VE TOT)

A calculated sum of the


volumes exhaled by the
patient for mandatory and
spontaneous breaths for the
previous one-minute interval
(also applies in BiLevel).

Range: 0.00 L/min to 99.9


L/min
Resolution: 0.01 L/min for
0.00 to 9.99 L/min; 0.1 L/min
for 10.0 to 99.9 L/min

Exhaled spontaneous minute


volume (VE SPONT)

The sum of exhaled spontaneous volumes per minute


(also applies in BiLevel)

Range: 0 L/min to 99.9 L/min


Resolution: 0.01 L/min for
0.00 to 9.99 L/min; 0.1 L/min
for 10.0 to 99.9 L/min

Exhaled spontaneous tidal


volume (VTE SPONT)

The exhaled volume of the


last spontaneous breath.

Range: 0 mL to 6000 mL
Resolution: 0.1mL for 0 mL to
9.9 mL; 1 mL for 10 mL to 6000
mL

Exhaled tidal volume (VTE)

The volume exhaled by the


patient for the previous mandatory or spontaneous breath
(also applies in BiLevel.

Range: 0 mL to 6000 mL
Resolution: 0.1mL for 0 mL to
9.9 mL; 1 mL for 10 mL to 6000
mL

Leak Sync exhaled tidal


volume (VTE)

The volume exhaled by the


patient for the previous mandatory or spontaneous breath
during Leak Sync (also applies
in BiLevel).

Range: 0 mL to 6000 mL
Resolution: 0.1mL for 0 mL to
9.9 mL; 1 mL for 10 mL to 6000
mL

I:E ratio

The ratio of the inspiratory


time to expiratory time for the
previous breath.

Range: 1:599 to 149:1


Resolution: 0.1 for 9.9:1 to
1:9.9; 1 for 149:1 to 10:1 and
1:10 to 1:599

Inspiratory compliance
(C20/C)

The ratio of compliance of the


last 20% of inspiration to the
compliance of the entire
inspiration

Range: 0 to 1.00
Resolution: 0.01

Intrinsic PEEP (PEEPI)

A calculated estimate of the


pressure above PEEP at the
end of exhalation.

Range: -20.0 cmH2O to +130


cmH2O
Resolution: 0.1 cmH2O
between -9.9 and +9.9
cmH2O; 1 cmH2O -10
cmH2O and 10 cmH2O

21

Specifications

Table11-11.Patient Data Range and Resolution (Continued)

22

Data Value

Description

Range and Resolution

Mean circuit pressure (PMEAN)

The calculated average circuit


pressure for an entire breath
cycle including both inspiratory and expiratory phases
(whether the breath is mandatory or spontaneous).

Range: -20.0 cmH2O to 100


cmH2O
Resolution: 0.1 cmH2O for 20.0 to 9.9 cmH2O; 1 cmH2O
for 10 to 100 cmH2O

Negative inspiratory force


(NIF)

The negative pressure generated during a maximally


forced inspiratory effort
against an obstruction to
flow.

Range: 0 cmH2O to -50


cmH2O
Resolution: 1 cmH2O for
values -10 cmH2O; 0.1
cmH2O for values > -10 cmH2O

O2% (monitored)

The monitored percentage of


oxygen in the gas delivered to
the patient, measured at the
ventilator outlet upstream of
the inspiratory filter.

Range: 0% to 103%
Resolution: 1%

P0.1

The inspiratory depression of


airway pressure after 100 ms
of occlusion. P0.1 is an indicator of respiratory drive.

Range: -20 cmH2O to 0


cmH2O
Resolution: 1 cmH2O when < 10 cmH2O; 0.1 cmH2O when
-10 cmH2O

PAV based intrinsic PEEP


(PEEPI PAV)

The estimated intrinsic PEEP


during a PAV+ breath. Intrinsic PEEP is an estimate of the
pressure above PEEP at the
end of every pause exhalation.

Range: 0 to 130 cmH2O


Resolution: 0.1 cmH2O for
values < 10 cmH2O; 1cmH2O
for values 10 cmH2O

PAV-based lung compliance


(CPAV)1

The calculated change in pulmonary volume for an applied


change in patient airway
pressure when measured
under conditions of zero flow
during a PAV+ plateau
maneuver. When PAV+ is
selected, the ventilator displays the current filtered
value for patient compliance,
and updates the display at the
successful completion of each
estimation. CPAV can be displayed in the vital patient data
banner. Reference Vital
Patient Data, p. 3-42.

Range: 2.5 mL/cmH2O to 200


mL/ cmH2O
Resolution: 0.1 mL/cmH2O for
values < 10 mL/cmH2O; 1
cmH2O for values 10 mL/
cmH2O

Operator's Manual

Performance Specifications

Table11-11.Patient Data Range and Resolution (Continued)


Data Value

Operator's Manual

Description

Range and Resolution

PAV-based lung elastance


(EPAV)1

For a PAV+ breath, EPAV is calculated as the inverse of CPAV


(see above). EPAV can be displayed in the vital patient data
banner. Reference Vital
Patient Data, p. 3-42.

Range: 5.0 cmH2O/L to 400


cmH2O/L
Resolution: 0.1 cmH2O/L for
values < 10 cmH2O/L; 1
cmH2O/L 10 cmH2O/L

PAV-based patient resistance


(RPAV)1

The difference between estimated total resistance RTOT


and the simultaneously estimated resistance of the artificial airway. When PAV+ is
selected, the ventilator displays the current filtered
value for patient resistance,
and updates the display at the
successful completion of each
estimation. RPAV can be displayed in the vital patient data
banner. Reference Vital
Patient Data, p. 3-42.

Range: 0.0 cmH2O/L/s to 60


cmH2O/L/s
Resolution: 0.1 cmH2O/L/s for
values < 10 cmH2O/L/s;
1cmH2O/ L/s for values 10
cmH2O/L/s

PAV-based total airway


resistance (RTOT)1

RTOT is an estimated value


captured just past peak expiratory flow and is equal to the
pressure loss across the
patient airway plus respiratory system (patient airway + ET
tube + expiratory limb of the
VBS)/ expiratory flow. This
pressure loss is divided by the
expiratory flow estimated at
the same moment, yielding
the estimate for RTOT.The
complete operation is orchestrated and monitored by a
software algorithm. When
PAV+ is selected, the ventilator displays the current filtered value for total
resistance, and updates the
display at the successful completion of each calculation.
RTOT can be displayed in the
vital patient data banner.
Reference Vital Patient Data,
p. 3-42.

Range: 1.0 cmH2O/L/s to 80


cmH2O/L/s
Resolution: 0.1 cmH2O/L/s for
values < 10 cmH2O/L/s; 1
cmH2O/ L/s for values 10
cmH2O/L/s

23

Specifications

Table11-11.Patient Data Range and Resolution (Continued)

24

Data Value

Description

Range and Resolution

PAV-based work of breathing


(WOBTOT)

The estimated effort needed


for patient inspiration including both patient and ventilator.

Range: 1.0 J/L to10.0 J/L


Resolution: 0.1 J/L

Peak expiratory flow (PEF)

The maximum speed of exhalation.

Range: 0 to 150 L/min


Resolution: 0.1 L/min for PEF
< 20 L/min; 1 L/min for PEF
20 L/ min

Peak circuit pressure (PPEAK)

The maximum pressure


during the previous breath,
relative to the patient wye,
including inspiratory and
expiratory phases.

Range: -20.0 cmH2O to 130


cmH2O
Resolution: 0.1 cmH2O for
values -20.0 cmH2O to 9.9
cmH2O; 1.0 cmH2O for values
10 cmH2O to 130 cmH2O

Peak spontaneous flow (PSF)

The maximum flow rate


sampled during a spontaneous inspiration.

Range: 0 to 200 L/min


Resolution: 0.1 L/min for
values < 20 L/min; 1L/min for
values 20 L/min

Plateau pressure (PPL)

The pressure measured


during an inspiratory pause
maneuver.

Range: -20.0 cmH2O to 130


cmH2O
Resolution: 0.1 cmH2O for
values -20.0 to 9.9 cmH2O; 1.0
cmH2O for values 10 cmH2O

Proximal exhaled tidal volume


(VTEY)

For neonatal patients, the


exhaled volume of the previous breath measured by the
Proximal Flow Sensor) (if
installed).

Range: 0 mL to 500 mL
Resolution: 0.1mL for values 0
mL to 9.9 mL; 1 mL for values
10 mL to 500 mL

Proximal exhaled total minute


volume (VE TOTY)

For neonatal patients, the


exhaled minute volume measured by the Proximal Flow
Sensor)

Range: 0.00 to 99.9 L/min


Resolution: 0.01 L/min for
0.00 to 9.99 L/min; 0.1 L/min
for 10.0 to 99.9 L/min

Proximal inspired tidal volume


(VTIY)

For neonatal patients, the


inspired volume of the previous breath measured by the
Proximal Flow Sensor) (if
installed).

Range: 0 mL to 500 mL
Resolution: 1 mL

Spontaneous inspiratory time


(TI SPONT)

The duration of the inspiratory phase of a spontaneous


breath.

Range: 0 s to 10 s
Resolution: 0.01 s

Operator's Manual

Performance Specifications

Table11-11.Patient Data Range and Resolution (Continued)

Operator's Manual

Data Value

Description

Range and Resolution

Spontaneous inspiratory time


ratio (TI/TTOT)

The fraction of the total spontaneous breath time used by


inspiration.

Range: 0 to 1
Resolution: 0.01

Spontaneous rapid shallow


breathing index (f/VT)

A calculated value using


exhaled spontaneous tidal
volume. High values indicate
the patient is breathing rapidly, but with little volume/
breath. Low values indicate
the inverse scenario.

Range: 0.1 1/min-L to 600


1/min- L
Resolution: 0.1 1/min-L for
values < 10 1/min-L; 1 1/min-L;
for values 10 1/min-L

Static compliance (CSTAT)

An estimate of the patients


lung-thorax static compliance
or elasticity.

Range: 0 mL/cmH2O to 500


mL/ cmH2O
Resolution: 0.1 mL/cmH2O for
values < 10 mL/; 1 mL/cmH2O
for values 10 mL/cmH2O

Resistance (RSTAT)

An estimate of the restrictiveness of the patients lungs


and the artificial airway.

Range: 0 cmH2O/L/s to 500


cmH2O /L/s
Resolution: 0.1 cmH2O/L/s for
values < 10 cmH2O/L/s, 1
cmH2O/ L/s for values 10
cmH2O/L/s

Total PEEP (PEEPTOT)

The estimated pressure at the


circuit wye during an expiratory pause maneuver.

Range: -20.0 cmH2O to +130


cmH2O
Resolution: 0.1 cmH2O for
values < 10 cmH2O; 1 cmH2O
for values -10 cmH2O and
10 cmH2O

Total respiratory rate (fTOT)

The number of mandatory or


spontaneous breaths/min
delivered to the patient.

Range: 1 to 200 1/min


Resolution: 0.1 1/min for
values < 10 1/min; 1 1/min for
10 1/min to 200 1/min

Vital capacity (VC)

The maximum amount of air


that can be exhaled after a
maximum inhalation.

Range: 0 mL to 6000 mL
Resolution: 0.1 mL for values
< 10 mL; 1 mL for values 10
mL

VLEAK

Inspiratory leak volume, the


total volume delivered during
inspiration to compensate for
the leak.

Range: 0 to 9000 mL
Resolution: 1 mL

25

Specifications

Table11-11.Patient Data Range and Resolution (Continued)


Data Value

Description

Range and Resolution

%LEAK

Percent leak, the percentage


of total delivered volume
during inspiration attributed
to the leak calculated as leak
volume during inspiration /
total delivered inspiratory
volume) x 100.

Range: 0 to 100%
Resolution: 1%

LEAK

Exhalation leak. The leak rate


at PEEP during exhalation.

Range: 0L/min to 200 L/min


Resolution: 0.1 L/min

LEAKY

Exhalation Leak at PEEP


during Leak Sync, measured
by the proximal flow sensor.

Range: 0 L/min to 200 L/min


Resolution: 0.1 L/min

1. If the estimated value of CPAV, EPAV, RPAV, or RTOT violates expected (PBW-based) limits, parentheses around the value indicate
the value is questionable. If the estimated value exceeds its absolute limit, the limit value flashes in parentheses.

Table11-12.Delivery Accuracy
Parameter

26

Accuracy

Range

Inspiratory pressure (PI)

(3.0 + 2.5% of setting) cmH2O

5 cmH2O to 90 cmH2O

End expiratory pressure (PEEP)

(2.0 + 4% of setting) cmH2O

0 cmH2O to 45 cmH2O

Pressure support (PSUPP)

(3.0 + 2.5% of setting) cmH2O

0 cmH2O to 70 cmH2O

Tidal volume (VT)

For adult and pediatric circuit type


settings:
For TI < 600ms:
(10 + 10% of setting x 600 ms/
TI ms) mL
For TI 600 ms
(10 + 10% of setting) mL
For neonatal circuit type settings:
For setting of 2 mL (VC+ only):
(1 + 10% of setting) mL
For setting of 3 mL to 4 mL:
(2+ 10% of setting) mL (delivered volume shall be 1 mL
For setting of 5 mL to 20 mL:
(3 + 15% of setting)
For setting of 20 mL:
(4 +10% of setting) mL

For adult and pediatric


circuit type settings:
25 mL to 2500 mL
For neonatal circuit type
settings: 2 mL to 310 mL

O2% (delivered)

3%

21% to 100%

PH

(2.0 + 4% of setting) cmH2O

5 cmH2O to 90 cmH2O

Operator's Manual

Performance Specifications

Table11-12.Delivery Accuracy (Continued)


Parameter

Accuracy

PL

(2.0 + 4% of setting) cmH2O

Range
0 cmH2O to 45cmH2O

Table11-13.Monitoring (Patient Data) Accuracy


Parameter

Operator's Manual

Accuracy

Range

Peak circuit pressure (PPEAK)

(2 + 4% of reading) cmH2O

5 cmH2O to 90 cmH2O

Mean circuit pressure (PMEAN)

(2 + 4% of reading) cmH2O

3 cmH2O to 70 cmH2O

End expiratory pressure (PEEP)

(2 + 4% of reading) cmH2O

0 cmH2O to 45cmH2O

End inspiratory pressure (PI END)

(2 + 4% of reading) cmH2O

5 cmH2O to 90 cmH2O

Inspired tidal volume (VTI)

(4 mL + 15% of actual) mL

2 mL to 2500 mL

Exhaled tidal volume (VTE)

(4 mL + 10% of actual) mL

2 mL to 2500 mL

Inspired tidal volume during


Leak Sync (VTL)

For adult and pediatric circuit


type settings:
For TI 600ms:
(10 + 20% x 600 ms/TI ms of
reading) mL
For TI > 600 ms:
(10 + 20% of reading) mL
For neonatal circuit type setting:
(10 + 20% of reading) mL
For readings < 100 mL, the
accuracy shall apply when the
percentage of inspiratory leak
volume is less than 80%

For adult and pediatric circuit


type settings:
25 mL to 2500 mL
For neonatal circuit type settings: 2 mL to 310 mL

Exhaled tidal volume (VTE)


during Leak Sync

For adult and pediatric circuit


type settings:
For TE 600ms:
(10 + 20% x 600 ms/TE ms of
reading) mL
For TE > 600 ms:
(10 + 20% of reading) mL
For neonatal circuit type setting:
(10 + 20% of reading) mL
For readings < 100 mL, the
accuracy shall apply when the
percentage of inspiratory leak
volume is less than 80%

For adult and pediatric circuit


type settings:
25 mL to 2500 mL
For neonatal circuit type settings: 2 mL to 310 mL

27

Specifications

Table11-13.Monitoring (Patient Data) Accuracy (Continued)


Parameter

Accuracy

Range

Proximal exhaled tidal volume


(VTEY)

(1 + 10% of reading) mL

2 mL to 310 mL

Proximal inspired tidal volume


(VTIY)

(1 + 10% of reading) mL

2 mL to 310 mL

O2% (monitored)

3%

15% to 100%

Respiratory rate (f)

0.8 1/min

1 1/min to 150 1/min

Table11-14.Computed Value Accuracy


Parameter

Accuracy

Range

PAV-based lung compliance (CPAV)

(1 + 20% of measured value)


mL/cmH2O

10 mL to 100 mL/cmH2O

PAV-based total airway


resistance (RTOT)

(3 + 20% of measured value)


cmH2O/L/s

5.0 cmH2O/L/s to 50 cmH2O/L/s

PAV-based work of
breathing (WOBTOT)

(0.5 + 10% of measured


work) J/L with a percent
support setting of 75%

0.7 J/L to 4 J/L

WARNING:
The ventilator accuracies listed in this chapter are applicable under the
operating conditions identified in the table Environmental Specifications on
Reference p. 11-7 .

Operation outside specified ranges cannot guarantee the accuracies listed in


the tables above, and may supply incorrect information.

11.8Regulatory Compliance
The ventilator complies with the following standards:

28

IEC 60601-1:2005 Medical Electrical Equipment, Part 1: General Requirements for


Basic safety and essential performance

EN 60601-1:2006, Medical Electrical Equipment, Part 1: General Requirements for


Basic safety and essential performance

Operator's Manual

Regulatory Compliance

ANSI-AAMI ES 60601-1:2005, Medical Electrical Equipment, Part 1: General


Requirements for Basic safety and essential performance

CSA C22.2 No. 60601-1:2008 Medical Electrical Equipment, Part 1: General


Requirements for Basic safety and essential performance

IEC 60601-1-8: 2006, Medical electrical equipment - Part 1-8: General requirements for basic safety and essential performance, Collateral Standard: General
requirements, tests and guidance for alarm systems in medical electrical equipment and medical electrical systems

EN 60601-1-8:2007, Medical electrical equipment - Part 1-8: General requirements for basic safety and essential performance, Collateral Standard: General
requirements, tests and guidance for alarm systems in medical electrical equipment and medical electrical systems

IEC 60601-2-12:2001, Medical electrical equipment Part 1-2: General requirements for basic safety and essential performance, Collateral standard: Safety
requirements for medical electrical systems

EN 60601-2-12:2005, Medical electrical equipment Part 2-12: Particular requirements for the safety of lung ventilators - Critical care ventilators

ISO/EN 80601-2-12: 2011, Medical electrical equipment Part 2-12: Particular


requirements for basic safety and essential performance of critical care ventilators

ISO 80601-2-55: 2011 and EN ISO 80601-2-55: 2012, Medical electrical equipment - Part 2-55: Particular requirements for the basic safety and essential performance of respiratory gas monitors - First Edition

ISO 5356-1:2004, Anesthetic and respiratory equipment Conical connectors Part


1: Cones and sockets

IEC 60601-1-4:2000, Medical Electrical Equipment - Part 1-4: General Requirements for Safety - Collateral Standard: Programmable Electrical Medical Systems

IEC 60601-1-6:2010, Medical electrical equipment - Part 1-6: General requirements for basic safety and essential performance - Collateral Standard: Usability

IEC 62366:2007, Medical devices - Application of usability engineering to medical


devices

ANSI/AAMI HE75:2009, Human Factors Engineering - Design of Medical Devices

IEC/EN 60601-1-2:2007, Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic compatibility - Requirements and tests

ISO 14971:2007/EN ISO 14971:2012, Medical devices - Application of risk management to medical devices

Operator's Manual

29

Specifications

11.9Manufacturers Declaration
The following tables contain the manufacturers declarations for the ventilator
system electromagnetic emissions, electromagnetic immunity, separation distances between ventilator and portable and mobile RF communications equipment and a list of compliant cables.

WARNING:
Portable and mobile RF communications equipment can affect the
performance of the ventilator system. Install and use this device according to
the information contained in this manual.
WARNING:
The ventilator system should not be used adjacent to or stacked with other
equipment, except as may be specified elsewhere in this manual. If adjacent
or stacked used is necessary, the ventilator system should be observed to
verify normal operation in the configurations in which it will be used.
Caution:
This equipment is not intended for use in residential environments and may
not provide adequate protection to radio communication services in such
environments.
Table11-15.Electromagnetic Emissions

The ventilator is intended for use in the electromagnetic environment specified below. The customer or the operator of the ventilator should assure that it is used in such an environment.
Emissions Test

30

Compliance

Electromagnetic environment guidance

Radiated RF emissions
CISPR 11

Group 1
Class A

The ventilator uses RF energy


only for its internal functions.
The ventilator is intended to be
used only in hospitals and not
be connected to the public
mains network.

Harmonic emissions IEC 61000-3-2

Class A

Voltage fluctuations/flicker
IEC 61000-3-3

Complies

The ventilator is intended to be


used only in hospitals and not
be connected to the public
mains network.

Operator's Manual

Manufacturers Declaration

Table11-16.Electromagnetic Immunity
The ventilator is intended for use in the electromagnetic environment specified below. The
customer or the operator of the ventilator should assure that it is used in such an environment.
Immunity Test

IEC 60601 test level

Compliance level

Electromagnetic
environment guidance

Electrostatic discharge
(ESD) IEC 61000-4-2

6 kV contact
8 kV air

6 kV contact
8 kV air

Floors should be
wood, concrete, or
ceramic tile. If floors
are covered with synthetic material, the relative humidity should
be at least 30%.

Electrical fast transient/


burst IEC 61000-4-4

2 kV for power
supply lines
1 kV for input/
output lines

2 kV for power
supply lines
1 kV for input/
output lines

Mains power quality


should be that of a
typical hospital environment.

Surge IEC 61000-4-5

1 kV line(s) to line(s)
2 kV line(s) to earth

1 kV line(s) to line(s)
2 kV line(s) to earth

Voltage dips, short


interruptions and
voltage variations on
power supply input
lines IEC/EN 61000-411

< 5% UT
(> 95% dip in UT) for
0.5 cycle
40% UT(60% dip in
UT) for 5 cycles
70% UT(30% dip in
UT) for 25 cycles
< 5% UT
> 95% dip in UT for 5 s

< 5% UT
(> 95% dip in UT) for
0.5 cycle
40% UT(60% dip in
UT) for 5 cycles
70% UT(30% dip in
UT) for 25 cycles
< 5% UT
> 95% dip in UT for 5 s

Mains power should


be that of a typical
hospital environment.
If the operator of the
ventilator requires
continuous operation
during power mains
interruptions, it is recommended that the
ventilator be powered
from an uninterruptible power supply or a
battery.

Power frequency
(50/60 Hz) magnetic
field
IE/EN 61000-4-8

3 A/m

3 A/m

Power frequency magnetic fields should be


at levels characteristic
of a typical hospital
environment.

NOTE: UT is the AC mains voltage prior to application of the test level.

Operator's Manual

31

Specifications

Table11-16.Electromagnetic Immunity (Continued)


The ventilator is intended for use in the electromagnetic environment specified below. The
customer or the operator of the ventilator should assure that it is used in such an environment.
Immunity Test

Conducted RF IEC/EN
61000-4-6

IEC 60601 test level

3 Vrms
150 kHz to 80 MHz

Compliance level

1 Vrms
159 kHz to 80 MHz

Electromagnetic
environment guidance
Portable and mobile
RF communications
equipment should be
used no closer to any
part of the ventilator
system, including
cables, than the separation distance calculated from the
equation applicable to
the frequency of the
transmitter.
Recommended separation distance

d = 3.5 P
10 Vrms in ISM bands1

1 Vrms in ISM bands1

Recommended separation distance

d = 12 P

32

Operator's Manual

Manufacturers Declaration

Table11-16.Electromagnetic Immunity (Continued)


The ventilator is intended for use in the electromagnetic environment specified below. The
customer or the operator of the ventilator should assure that it is used in such an environment.
Immunity Test

Radiated RF IEC/EN
61000-4-3

IEC 60601 test level

10 V/m
80 MHz to 2.5 GHz

Compliance level

Electromagnetic
environment guidance

10 V/m Modulation of
80% AM @ 2 Hz
80 MHz to 2.5 GHz

d = 1.2 P

80 MHz to 800 MHz

d = 2.3 P

800 MHz to 2.5 GHz


Where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer and d is the separation distance in meters (m)2. Field strengths from fixed transmitters, as determined by an electromagnetic site survey3, should be less than the compliance level in
each frequency range4. Interference may occur in the vicinity of equipment marked with the following
symbol:

NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies
NOTE 2 these guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.
1. The ISM (industrial, scientific and medical) bands between 150 kHz and 80 MHz are 6.765 to 6,795 MHz; 13.553 MHz to
13.567 MHz; 26.957 MHz; and 40.66 MHz to 40.70 MHz. The compliance levels in the ISM frequency bands between 150 kHz
and 80 MHz and in the frequency range 80 MHz to 2.5 GHz are intended to decrease the likelihood mobile/portable communications equipment could cause interference if it is inadvertently brought into patient areas. For this reason, an additional factor of
10/3 is used in calculating the separation distance for transmitters in these frequency ranges.

Operator's Manual

33

Specifications

2. The compliance levels in the ISM frequency bands between 150 kHz and 80 MHz and in the frequency range 80 MHz to 2.5 GHz
are intended to decrease the likelihood mobile/portable communications equipment could cause interference if it is inadvertently
brought into patient areas. For this reason, an additional factor of 10/3 is used in calculating the separation distance for transmitters
in these frequency ranges.
3. Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur
radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic
environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in
the location in which the 980 Series Ventilator is used exceeds the applicable RF compliance level above, the 980 Series Ventilator
should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such
as reorienting or relocating the ventilator.
4. Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 10 V/m.

Table11-17.Recommended Separation Distances for RF


The ventilator is intended for use in an electromagnetic environment in which radiated RF
disturbances are controlled. The customer or the operator of the ventilator can help prevent
electromagnetic interference by maintaining a minimum distance between portable and
mobile RF communications equipment (transmitters) and the ventilator as recommended
below, according to the maximum output power of the communications equipment.
Rated
maximum
output power
of transmitter
(W)

150 kHz to 80
MHz outside of
ISM bands

150 kHz to 80
MHz inside of
ISM bands

d = 3.5 P

d = 12 P

80MHz to 800
MHz

800 MHz to 2.5


GHz

d = 1.2 P

d = 2.3 P

0.01

0.35

1.2

0.12

0.23

0.1

1.1

3.8

0.38

0.73

3.5

12

1.2

2.3

10

11

38

3.8

7.3

100

35

120

12

23

For transmitters rated at a maximum output power not listed above, the recommended separation
distance d in meters (m) can be estimated using the equation applicable to the frequency of the transmitter where P is the maximum output power rating of the transmitter in watts (W) according to the
transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.

34

WARNING:
The use of accessories and cables other than those specified with the
exception of parts sold by Covidien as replacements for internal components,
may result in increased emissions or decreased immunity of the ventilator
system.

Operator's Manual

Safety Tests

Table11-18.
Part number and description

Cable length

10087151, Power cord, 10A, RA, ANZ

10 ft (3 m)

10087159, Power cord, 10A, RA, UK

10 ft (3 m)

10087155, Power cord, 10A, RA, EU

10 ft (3 m)

10087157, Power cord, 10A, RA, Japan

10 ft (3 m)

10087152, Power cord, 10A, RA, British

10 ft (3 m)

10087154, Power cord, 10A, RA, Swtzrlnd

10 ft (3 m)

10081056, Power cord, 10A, RA, USA

10 ft (3 m)

10087156, Power cord, 10A, RA, Israel

10 ft (3 m)

10087160, Power cord, 10A, RA, Brazil

10 ft (3 m)

10087153, Power cord, 10A, RA, China

10 ft (3 m)

11.10Safety Tests
All safety tests should be performed by qualified Service personnel at the interval specified. Reference Service Preventive Maintenance Frequency, p. 1-2 in
the Operators Manual Addendum on Preventive Maintenance.

11.11Essential Performance Requirements


Per ISO/EN 80601-2-12: 2011, Medical electrical equipment Part 2-12: Particular requirements for basic safety and essential performance of critical care
ventilators, the ventilators essential performance requirements are given in
Ventilator Settings, Alarm Settings, and Patient Data tables earlier in this chapter. Alarms, including Oxygen level alarms and gas failure alarms, are identified
in Chapter 6. AC and battery backup power information is given in Chapter 3,
and gas failure cross flow information is given in Chapter 3.

Operator's Manual

35

Specifications

Page Left Intentionally Blank

36

Operator's Manual

A BiLevel 2.0 Appendix

A.1

Overview
This appendix describes the operation of the BiLevel 2.0 ventilation mode on
the Puritan Bennett 980 Series Ventilator.
BiLevel is a mixed mode of ventilation that combines attributes of mandatory
and spontaneous breathing, with the breath timing settings determining
which breath type is favored. In BiLevel Mode, mandatory breaths are always
pressure-controlled, and spontaneous breaths can be pressure-supported (PS)
or tube compensated (TC).
FigureA-1.Spontaneous Breathing at PL

PCIRC (cmH2O)

PH

TH

PL

TL

Spontaneous breaths

BiLevel resembles SIMV mode, except that BiLevel establishes two levels of positive airway pressure. Cycling between the two levels can be triggered by
BiLevel timing settings or by patient effort.

A-1

BiLevel 2.0 Appendix

FigureA-2.BiLevel Mode

Pressure (y-axis)

Synchronized transitions

PL

Pressure support

PH

Time-based transitions

Spontaneous breath

NOTE: The Pressure support level is


always referenced to PL

The two pressure levels are called Low Pressure (PL) and High Pressure (PH). At
either pressure level, patients can breathe spontaneously, and spontaneous
breaths can be assisted with tube compensation or pressure support. BiLevel
monitors mandatory and spontaneous tidal volumes separately.
Inspiratory time and expiratory time in BiLevel become Time high (TH) and Time
low (TL), respectively. During these inspiratory and expiratory times, PH is maintained during TH and PL is maintained during TL.
A.2

Intended Use
BiLevel is intended for adult, pediatric, and neonatal patients.

A.3

Safety Symbol Definitions


This section contains safety information for users, who should always exercise
appropriate caution while using the ventilator.

A-2

Operators Manual

Technical Description

TableA-1.Safety Symbol Definitions


Symbol

Definition
WARNING
Warnings alert users to potential serious outcomes (death, injury, or adverse
events) to the patient, user, or environment.
Caution
Cautions alert users to exercise appropriate care for safe and effective use of the
product.
Note
Notes provide additional guidelines or information.

A.4

WARNING:
The ventilator offers a variety of breath delivery options. Throughout the
patient's treatment, the clinician should carefully select the ventilation mode
and settings to use for that patient based on clinical judgment, the condition
and needs of the patient, and the benefits, limitations and characteristics of
the breath delivery options. As the patient's condition changes over time,
periodically assess the chosen modes and settings to determine whether or
not those are best for the patient's current needs.

Setting Up BiLevel
BiLevel is a ventilatory mode (along with A/C, SIMV, and SPONT).
To set up BiLevel
1.
At the ventilator setup screen, enter PBW or gender and height.
2.

Touch BiLevel. After selecting BiLevel mode, the ventilator uses the PC mandatory
breath type, which cannot be changed.

3.

Choose the spontaneous type (PS or TC).

4.

Choose trigger type (P-TRIG or V-TRIG).

5.

Operators Manual

Select desired ventilator settings. The default settings for BiLevel mode appear. To
change a setting, touch its button and turn the knob to set its value. PH must
always be at least 5 cmH2O greater than PL.

A-3

BiLevel 2.0 Appendix

Note:
The rise time % setting determines the rise time to reach target pressure for
transitions from PL to PH and for spontaneous breaths, even when pressure support
(PSUPP) = 0. Expiratory sensitivity (ESENS) applies to all spontaneous breaths.
6.

Set TL, TH, or the ratio of TH to TL. To select settings that would result in a TH:TL
ratio greater than 1:1 or 4:1, you must touch Continue to confirm after reaching
the 1:1 and 4:1 limits.
FigureA-3.BiLevel Setup Screen

A.5

7.

Touch Start.

8.

Set apnea and alarm settings by touching their respective tabs at the side of the
ventilator settings screen and changing settings appropriately.

Using Pressure Support with BiLevel


Spontaneous breaths in BiLevel mode can be assisted with pressure support
according to these rules (Reference BiLevel with Pressure Support, p. A-5):

A-4

Pressure support (PSUPP) can be used to assist spontaneous breaths at PL and PH.
PSUPP is always set relative to PL. Target pressure = PL + PSUPP.

Operators Manual

Technical Description

Spontaneous patient efforts at PH are not pressure supported unless PSUPP > (PH PL). All spontaneous breaths (whether or not they are pressure supported) are
assisted by a pressure of 1.5 cmH2O.

If PSUPP + PL is greater than PH + 1.5 cmH2O, all spontaneous breaths at PL are


assisted by the PSUPP setting, and all spontaneous breaths at PH are assisted by
PSUPP - (PH - PL).

All spontaneous breaths not supported by PS or TC (for example, a classic CPAP


breath) are assisted with an inspiratory pressure of 1.5 cmH2O.

For example, if PL = 5 cmH2O, PH = 15 cmH2O, and PSUPP = 20 cmH2O:

All spontaneous breaths at PL are assisted by 20 cmH2O of pressure support (PL +


PSUPP) for a total pressure of 25 cmH2O, and

All spontaneous breaths in PH are assisted by 10 cmH2O of pressure support


(PSUPP - (PH - PL)) for the same total pressure of 25 cmH2O.
FigureA-4.BiLevel with Pressure Support

Pressure (y-axis)

PH

PH Pressure support = 10 cmH2O

PL

PL Pressure support = 20 cmH2O

During spontaneous breaths, the pressure target is calculated with respect to PL.
A.6

Manual Inspirations in BiLevel Mode


Pressing the MANUAL INSP key during BiLevel mode causes the ventilator to:

Operators Manual

A-5

BiLevel 2.0 Appendix

Cycle to PH, if the current pressure level is PL.

Cycle to PL, If the current pressure level is PH.

To avoid breath stacking, the ventilator does not cycle from one pressure level
to another during the earliest stage of exhalation.
A.7

Respiratory Mechanics Maneuvers in BiLevel


In BiLevel, respiratory mechanics maneuvers are limited to inspiratory pause
and expiratory pause maneuvers.

A.8

Specifications
Reference the table, Ventilator Settings Range and Resolution, in Chapter 11
of this manual for the following specifications:

A.9

Low pressure (PL)

High pressure (PH)

Low pressure time (TL)

High pressure time (TH)

TH:TL ratio

Respiratory rate (f)

Rise time %

Technical Description
BiLevel is a mode of ventilation that alternately cycles between two operatorset pressure levels, PL and PH. The pressure durations are defined by operatorset timing variables TL and TH. Transitions between the two pressure levels, PL
and PH, are analogous to breath phase transitions in PC.
At the extreme ranges of TL and TH, BiLevel can resemble the single breath type
mode A/C - PC, or the more complex breath type mode, an inverted-like
IMV. If TH and TL assume normal values with respect to PBW (for example

A-6

Operators Manual

Technical Description

TH:TL 1:2 or 1:3), then BiLevel assumes a breathing pattern similar to, if not
qualitatively identical to A/C - PC. However, as TL begins to shorten with the
TH:TL ratio extending beyond 4:1, the breathing pattern assumes a distinctly
different shape. In the extreme, the exaggerated time at PH and abrupt release
to PLwould match the pattern patented by John Downs* and defined as APRV.
In between the A/C-PC-like pattern and the APRV-like pattern, there would be
patterns with moderately long TH and TL intervals, allowing the patient sufficient time to breathe spontaneously at both PH and PL. In these types of
breathing patterns, (but less so with APRV) BiLevel, like SIMV, can be thought
of as providing both mandatory and spontaneous breath types. In this sense,
BiLevel and SIMV are classified as mixed modes.
Direct access to any of the three breath timing parameters in BiLevel is accomplished by touching the Padlock icon associated with the TH period, TL period
or the TH:TL ratio displayed on the breath timing bar in the setup screen.
While in BiLevel mode, spontaneously triggered breaths at either pressure level
can be augmented with higher inspiratory pressures using Pressure Support
(PS) or Tube Compensation (TC) breath types.
A.9.1

Synchrony in BiLevel
Just as BiLevel attempts to synchronize spontaneous breath delivery with the
patient's inspiratory and expiratory efforts, it also attempts to synchronize the
transitions between pressure levels with the patient's breathing efforts. This
allows TH to be extended to prevent transitions to PL during the patient's spontaneous inspiration. Likewise, the TL interval may be extended to prevent a
transition to PH during the patient's spontaneous exhalation.
The trigger sensitivity setting (PSENS or VSENS) is used to synchronize the transition from PL to PH. The transition from PH down to PL is synchronized with the
patient's spontaneous expiratory effort. The BiLevel algorithm will vary the TL
and TH intervals as necessary to synchronize the transitions between PL and PH
to match the patient's breathing pattern.
The actual durations of TH and TL vary according to whether or not the patient
makes any spontaneous inspiratory efforts during those periods.

*. Downs, JB, Stock MC. Airway pressure release ventilation: A new concept in ventilatory support. Crit Care Med 1987;15:459-461

Operators Manual

A-7

BiLevel 2.0 Appendix

To manage synchrony with the patient's breathing pattern, the BiLevel algorithm partitions the TH and TL periods into spontaneous and synchronous intervals as shown in the figure below.
FigureA-5.Spontaneous and Synchronous Intervals

Pressure (y-axis)

TL

TH

Synchronous interval

PH

Spontaneous interval

PL

By partitioning TH and TL into spontaneous and synchronous phases, BiLevel


responds to patient efforts (or lack of them) in a predictable pattern:

A.9.2

During the spontaneous interval of each pressure level, successful inspiratory


efforts cause the ventilator to deliver spontaneous breaths.

During TL synchronous intervals, successful inspiratory efforts cause the ventilator


to cycle from PL to PH. If there is no spontaneous (patient) effort, this transition
takes place at the end of the TL period.

During TH synchronous intervals, successful expiratory efforts cause the ventilator


to cycle from PH to PL. If there is no spontaneous exhalation, the transition to the
PL level takes place at the end of the TH period.

Patient Monitoring in BiLevel


If the patient breathes spontaneously at either pressure level, BiLevel monitors
and displays the total respiratory rate, including mandatory and spontaneous
breaths. BiLevel also displays the exhaled tidal volume and total exhaled minute

A-8

Operators Manual

Technical Description

volume for both mandatory and spontaneous breaths.


A.9.3

APRV Strategy in BiLevel


Lengthening the TH period and shortening the TL period to only allow incomplete exhalation of the mandatory breath volume, results in an inverse TH:TL
ratio. In this breath timing configuration with TH:TL ratios of greater than 4:1,
BiLevel becomes Airway Pressure Release Ventilation (APRV).
APRV is characterized as longer TH periods, short TL periods (usually less than
one second), and inverse TH:TL ratios. Since, at these breath timing settings, all
of the patient-triggered spontaneous breaths occur during the TH period,
APRV resembles CPAP ventilation with occasional, short periods of incomplete
exhalation referred to as releases which are controlled by the f setting.
FigureA-6.APRV With Spontaneous Breathing at PH

PCIRC (cmH2O)

Lengthened inspiratory time (TH)

Shortened release time (TL)

In APRV, the PH level is set to optimize pulmonary compliance for spontaneous


breathing while maintaining an elevated mean airway pressure to promote
oxygenation, the PL level is set, along with the TL, to control the expiratory
release volume of mandatory breaths to help manage CO2 and alveolar ventilation, and the f setting controls the number of releases per minute which are
used to help manage the patient's CO2 levels. The f setting also impacts the
mean airway pressure.

Operators Manual

A-9

BiLevel 2.0 Appendix

In APRV the operator can configure the BiLevel settings to allow direct control
of TL to assure that changes in the f setting will not inadvertently lengthen the
TL period resulting in destabilization of end-expiratory alveolar volume. With
the TL period locked, changes in set f will change the TH period to accommodate the new f setting while maintaining the set TL period.
A.9.4

Technical Structure of BiLevel


In BiLevel, the ventilator establishes two levels of baseline pressure. One level
is essentially the same as the standard PEEP level set for all common modes of
ventilation. The second pressure level is the level established at TH. Both pressure levels permit CPAP, TC and PS breaths. The breath timing settings determine whether the patient can initiate any of these breath types.

A.10

Mode Changes
Changing to BiLevel mode from other modes follows the general guidelines for
mode changes:

A-10

The change is made as soon as possible without compromising inspiration or


exhalation.

Breaths are not stacked during inspiration.

Operators Manual

B Leak Sync Appendix

B.1

Overview
This appendix describes the operation of the Puritan Bennett 980 Series Ventilator Leak Sync option. The Leak Sync option enables the ventilator to compensate for leaks in the breathing circuit while accurately detecting the
patients effort to trigger and cycle a breath. Because Leak Sync allows the ventilator to differentiate between flow due to leaks and flow due to patient respiratory effort, it provides dynamic compensation and enhances patientventilator synchrony. Reference Chapter 4 in this manual for general parameter and operational information.

B.2

Intended Use
Leak Sync is designed to compensate for leaks in the breathing circuit during
noninvasive or invasive ventilation. Leak Sync accurately quantifies instantaneous leak rates, therefore detecting patient respiratory phase transitions correctly and may affect work of breathing. Leak Sync is intended for neonatal,
pediatric, and adult patients.

B-1

Leak Sync Appendix

B.3

Safety Symbol Definitions


This section contains safety information for users, who should always exercise
appropriate caution while using the ventilator.
TableB-1.Safety Symbol Definitions
Symbol

Definition
WARNING
Warnings alert users to potential serious outcomes (death, injury, or adverse
events) to the patient, user, or environment.
Caution
Cautions alert users to exercise appropriate care for safe and effective use of the
product.
Note
Notes provide additional guidelines or information.

B.4

WARNING:
The ventilator offers a variety of breath delivery options. Throughout the
patient's treatment, the clinician should carefully select the ventilation mode
and settings to use for that patient based on clinical judgment, the condition
and needs of the patient, and the benefits, limitations and characteristics of
the breath delivery options. As the patient's condition changes over time,
periodically assess the chosen modes and settings to determine whether or
not those are best for the patient's current needs.

Leak Sync
Breathing circuit leaks can cause the ventilator to erroneously detect patient
inspiratory efforts (called autotriggering) or delay exhalation in pressure support. Patient interfaces such as masks are particularly prone to significant leaks.
Inaccurately declaring inspiration or exhalation can result in patient-ventilator
dysynchrony and increased work of breathing.
Changing inspiratory or expiratory sensitivity settings can temporarily correct
the problem, but requires continued frequent clinical intervention to ensure
that sensitivity is adjusted appropriately as conditions change (for example, if
the patient moves or the circuit leak changes).

B-2

Operators Manual

Technical Discussion

Leak Sync adds flow to the breathing circuit to compensate for leaks. The
maximum Leak Sync flow applies to the maximum base flow compensation
during exhalation. During pressure-based inspirations, the total delivered flow
(leak flow plus inspiratory flow) is limited by the maximum total flow.
The following table shows the maximum leak rates at set PEEP pressure that
Leak Sync will compensate based on patient type.
TableB-2.Leak Compensation Volumes Based on Patient Type
Patient type

B.5

Maximum Leak compensation flow

Neonatal

15 L/min

Pediatric

40 L/min

Adult

65 L/min

Setting Up Leak Sync


For more information on setting up the ventilator, Reference Chapter 4 of this
manual.
To enable Leak Sync
1.
At the ventilator setup screen, touch the More Settings tab.
2.

Touch Enabled in the Leak Sync area.

3.

Touch Accept ALL to enable Leak Sync.

Operators Manual

B-3

Leak Sync Appendix

FigureB-1.Enabling Leak Sync

B.6

Note:
The default value for Leak Sync is Disabled when the circuit type is Pediatric or Adult
and the Vent Type is Invasive. Otherwise the default value for Leak Sync is Enabled.
Note:
Leak Sync is not allowed for tube compensated (TC) and Proportional Assist
Ventilation (PAV+) breath types.

When Leak Sync is Enabled


Reference GUI Screen when Leak Sync is Enabled, p. B-5 for an example
showing the GUI screen when Leak Sync is enabled.

B-4

The Vent Setup button on the GUI screen indicates Leak Sync is active.

DSENS is displayed in units of L/min, rather than %.

If the ventilator detects a leak during a respiratory mechanics maneuver, the


message Leak Detected is displayed.

A new leak or change in leak rate is typically quantified and compensated within
three breaths. Monitored patient data stabilizes within a few breaths.

Operators Manual

Technical Discussion

Select inspiratory and expiratory sensitivity settings as usual. If the ventilator autotriggers, try increasing flow sensitivity (VSENS)

Note:
The absence of the Leak Detected message does not mean there is no leak.
Note:
Leak Sync is automatically enabled when ventilating a new patient and the circuit type
is neonatal, regardless of the Vent Type. If Leak Sync is disabled, it remains disabled
when switching between INVASIVE and NIV Vent Types.
FigureB-2.GUI Screen when Leak Sync is Enabled

B.6.1

LS appears on Vent Setup button notifying the operator that Leak Sync is enabled

Adjusting Disconnect Sensitivity (DSENS)


When Leak Sync is enabled, the Circuit Disconnect alarm becomes active based
on the DSENS setting, which is the maximum allowable leak rate at set PEEP.
When Leak Sync is disabled, DSENS is automatically set to 75%.

Operators Manual

B-5

Leak Sync Appendix

WARNING:
When Vent Type = NIV and Leak Sync is disabled, DSENS is automatically set to
OFF.

Reference the table below for a summary of DSENS settings when Leak Sync is
enabled. Note that it is possible to set DSENS below maximum Leak Sync flow.
TableB-3.DSENS settings

B.6.2

Breathing circuit type

DSENS setting

Maximum total flow

Neonatal

Range: 1 to 15 L/min
Default: 2 L/min (INVASIVE ventilation) 5 L/min (NIV)

50 L/min

Pediatric

Range: 1 to 40 L/min
Default: 20 L/min

120 L/min

Adult

Range: 1 to 65 L/min
Default: 40 L/min

200 L/min

WARNING:
With significant leaks, pressure targets may not be reached due to flow
limitations.
WARNING:
Setting DSENS higher than necessary may prevent timely detection of
inadvertent extubation.

Monitored Patient Data


When Leak Sync is enabled, three additional parameters are displayed on the
More Patient Data screen and updated for each breath. Display the More
Patient Data screen by swiping the tab on the patient data banner. These leak
parameters may also be configured on the patient data banner and the large
font patient data panel.

B-6

Operators Manual

Technical Discussion

FigureB-3.Leak Sync Monitored Patient Data

Leak Sync Parameters

Reference the table Patient Data Range and Resolution in Chapter 11 of this
manual for information regarding the following monitored patient data
parameters:

VLEAK

% LEAK

LEAK

Displayed values for Exhaled Tidal Volume (VTE ) and Inspired Tidal Volume
(VTL) are leak-compensated, and indicate the estimated inspired or exhaled
lung volume. The accuracies for VTE and VTL also change when Leak Sync is
enabled (see Technical Discussion for more information). Graphic displays of
flow during Leak Sync indicate estimated lung flows.

Operators Manual

B-7

Leak Sync Appendix

B.7

Technical Discussion
Managing breathing circuit leaks is important to ensure appropriate breath
triggering and cycling, ventilation adequacy, and valid patient data. Detecting
and monitoring leaks can improve treatment, reduce patient work of breathing, and provide more accurate information for clinical assessments.
Leak Sync recognizes that changing pressures lead to varying deflection of
interface materials and leak sizes. The Leak Sync leak model includes a rigid
leak orifice whose size remains constant under changing pressures, combined
with an elastic leak source whose size varies as a function of applied pressure.
This algorithm provides a more accurate estimate of instantaneous leak to
improve patient-ventilator synchrony under varying airway pressures.
Leak Sync allows the ventilator to determine the leak level and allows the operator to set the flow trigger and peak flow sensitivities to a selected threshold.
The base flow during exhalation is set to:

B.7.1

Flow triggering: 1.5 L/min + estimated leak flow at PEEP + flow sensitivity.

Pressure triggering: 1.0 L/min + estimated leak flow at PEEP.

Inspired Tidal Volume (VTL) Accuracy During Leak Sync


Reference Patient Data Range and Resolution, p. 11-19, VTL parameter, for VTL
accuracy.
For readings < 100 mL, accuracy ranges apply when the percentage of inspiratory leak volume is < 80%, where the percentage of leak volume is:
(Leak volume during inspiration / total delivered inspiratory volume) x 100

B.7.2

Note:
Inspired tidal volume is labeled as VTL when Leak Sync is enabled, and as VTI when
Leak Sync is disabled.

Exhaled Tidal Volume (VTE) Accuracy During Leak Sync


Reference Patient Data Range and Resolution, p. 11-19, VTE parameter, for
accuracy when Leak Sync is enabled.
where TE = time to exhale 90% of volume actually exhaled by the patient.

B-8

Operators Manual

Technical Discussion

For readings < 100 mL, accuracy ranges apply when the percentage of inspiratory leak volume is < 80%, where the percentage of leak volume is:
(Leak volume during inspiration/total delivered inspiratory volume) x 100
B.7.3

%LEAK Calculation
Reference Patient Data Range and Resolution, p. 11-19, % LEAK parameter,
for specifications.

B.7.4

Circuit Disconnect Alarm During Leak Sync


The Circuit Disconnect alarm is activated if the overall leak volume during the
whole breath exceeds the maximum leak volume derived from the DSENS setting. During VC, the Circuit Disconnect alarm is also activated if the end-inspiratory pressure falls below (set PEEP + 1 cmH2O) for three consecutive breaths.
The screen shows this alarm message:
FigureB-4.Circuit Disconnect During VC

Normal operation resumes if the ventilator detects a patient connection.

Operators Manual

B-9

Leak Sync Appendix

Page Left Intentionally Blank

B-10

Operators Manual

C IE Sync Appendix

C.1

Overview
This appendix describes the operation of the Puritan Bennett 980 Ventilator
IE Sync option.
When a mechanically ventilated patient is capable of some degree of ventilatory effort, it is essential that the ventilator be as responsive as possible to the
patients efforts. In most modes and breath types, this is done by clinicians estimating PSENS, VSENS, and ESENS.
IE Sync is a ventilator triggering and cycling algorithm based on real-time estimation of the patients intra-pleural pressure. The algorithm processes ventilator pressure and flow measurements in a way that accounts for the variability
of respiratory parameters of the patient population. In a bench-top model simulating patients with flow limitation, incomplete exhalation or weak spontaneous efforts, IE Sync triggering has been shown to decrease the number of
missed breaths when compared to flow triggering.

C.2

Intended Use
IE Sync is intended for invasively ventilated, spontaneously breathing pediatric
and adult patients, only, and not for neonatal patients.

C.3

Safety Symbol Definitions


This section contains safety information for users, who should always exercise
appropriate caution while using the ventilator.

C-1

IE Sync Appendix

TableC-1.Safety Symbol Definitions


Symbol

Definition
WARNING
Warnings alert users to potential serious outcomes (death, injury, or adverse events) to
the patient, user, or environment.
Caution
Cautions alert users to exercise appropriate care for safe and effective use of the
product.
Note
Notes provide additional guidelines or information.

C.4

WARNING:
The ventilator offers a variety of breath delivery options. Throughout the
patient's treatment, the clinician should carefully select the ventilation mode
and settings to use for that patient based on clinical judgment, the condition
and needs of the patient, and the benefits, limitations and characteristics of
the breath delivery options. As the patient's condition changes over time,
periodically assess the chosen modes and settings to determine whether or
not those are best for the patient's current needs.

IE Sync Description
IE Sync is a triggering and cycling algorithm based on the estimated patient
intra-pleural pressure. It is only available in SPONT mode, with the following
breath types: PAV+, PS, VS, and TC. A PBW 25 kg is recommended. IE Sync
can also be used in monitoring mode, where it does not control the patient's
breath triggering and cycling, but only indicates where in the patient's breath
cycle an inspiration and expiration would have started if IE Sync were active.

C.5

Setting Up IE Sync
To set up IE Sync
1.
At the ventilator setup screen, enter the patients gender and height or PBW.
2.

C-2

Touch the Invasive vent type button.

Operators Manual

Technical Description

3.

Touch SPONT mode.

4.

Touch PAV+, VS, PS, or TC to choose the spontaneous type.

5.

Touch IE Sync as the trigger type.

6.

Set the inspiratory triggering threshold (ISENS).

7.

Set the expiratory sensitivity threshold (ESENS).

8.

Continue setting up the ventilator as described in Chapter 4 of this manual.


FigureC-1.Setup Screen

Note:
If Leak Sync is currently enabled, it becomes disabled when IE Sync is selected as
the trigger type. Leak Sync cannot be selected if using IE Sync as the trigger type.
Conversely, If IE Sync is disabled, then Leak Sync becomes selectable.

A message appears on the GUI stating IE Sync Startup until the IE Sync
startup period completes. The startup period takes three (3) breaths to reach a
steady state, after which it triggers and cycles the ventilator breaths. During
the startup period, the default ventilator settings are:

Trigger type P-TRIG

PSENS 2 cmH2O

Operators Manual

C-3

IE Sync Appendix

ESENS 25% if IE Sync has not been used previously, otherwise default ESENS
setting for the currently active breath type

If IE Sync is enabled while in transition from a different trigger type, then the
trigger type, PSENS, and ESENS settings from the prior breath type shall be used
during startup instead of those stated above.
C.6

Setting Up IE Sync for Monitoring Only


To set up IE Sync for monitoring only:
1.
At the Vent Setup screen, change the trigger type to a type other than IE Sync.
2.

Touch Accept.

3.

Swipe the Menu tab on the left side of the GUI.


FigureC-2.Monitoring Setup

4.

C-4

Touch IE Sync. A dialog appears allowing enabling (ON) or disabling (OFF) of IE


Sync monitoring.

Operators Manual

Technical Description

FigureC-3.Monitoring Dialog

C.7

5.

Touch ON.

6.

Set the inspiratory trigger threshold (ISENS).

7.

Set the expiratory sensitivity threshold (ESENS).

8.

Touch Close.

IE Sync Monitoring Display


The waveforms screen shows the PSYNC (intra-pleural pressure) signal and the
I:E signal during the delivery of a spontaneous breath. Reference Technical
Description, p.C-7 for details regarding PSYNC.

Note:
IE Sync-related signals are not displayed during Apnea ventilation.

IE Sync monitoring displays a waveform indicator marking where in the breath


cycle the ventilator would have triggered an inspiration or expiration and
started an expiration if IE Sync were in use.

Operators Manual

C-5

IE Sync Appendix

When using IE Sync monitoring, the waveform displays an indication of missed


breath triggers, if applicable (missed triggers per minute and the percent of
missed triggers).

Note:
IE Sync-related signals are not displayed during Apnea ventilation.

When using IE Sync monitoring, the screen appears as shown below.


FigureC-4.Monitoring Waveform

The points at which the actual inspiration and expiration took place under pressure trigger or flow
trigger

I:E signal: The points in the breath where IE Sync would have triggered an inspiration and expiration
if not in monitoring

C-6

Note:
In the PCIRC waveform above, red indicates the current breath trigger and cycle timing,
and blue indicates where IE Sync would have triggered inspiration and exhalation, had
IE Sync been the chosen trigger type.

Operators Manual

Technical Description

C.8

Technical Description
IE Sync uses a real-time estimate,
(PSYNC), of the time-derivative of intrapleural pressure for determining breath phase transitions. PSYNC is computed
every five (5) ms based on a proprietary algorithm that uses the ventilators
pressure and flow sensor readings as its inputs. IE Sync detects the start of
inspiration when PSYNC reaches the set ISENS threshold, and detects the end of
inspiration when the PSYNC signal reaches the set ESENS threshold.

C.9

Alarms during IE Sync


If an alarm occurs during IE Sync which causes the ventilator to enter inspiratory
or expiratory Back Up Ventilation (BUV), the trigger type automatically changes
to the factory default trigger type, which can be institutionally configured. If and
when the alarm that caused BUV is corrected, IE Sync will not automatically be
re-enabled; if IE Sync is still desired, then the user must re-enable it.
An apnea condition during IE Sync causes the trigger type to change to P-TRIG
with PSENS of two (2) cmH2O.
IE Sync breaths go through a startup period when normal ventilation resumes
after the following resets:

Circuit occlusion

Circuit disconnect

Loss of both gas supplies (SVO)

Calculated lung compliance value is invalid

Calculated patient resistance value is invalid

If there are five (5) internal resets in a 20-minute period, IE Sync will automatically become disabled, and will be set to V-TRIG with new-patient default sensitivities. Reference the table Ventilator Settings Range and Resolution in
Chapter 11 of this manual. A message displays on the GUI if IE Sync becomes
disabled due to internal resets, and the unread items icon appears on the ventilator setup dialog.

Operators Manual

C-7

IE Sync Appendix

Page Left Intentionally Blank

C-4

Operators Manual

D PAV+ Appendix

D.1

Overview
This appendix describes the operation of the PAV*+ software option for the
Puritan Bennett 980 Ventilator.
Proportional Assist* Ventilation (PAV+) is designed to improve the work of
breathing of a spontaneously breathing patient by reducing the patients
increased work of breathing when pulmonary mechanics are compromised.
The PAV+ breath type differs from the pressure support (PS) breath type in the
following way:
PAV+ acts as an inspiratory amplifier; the degree of amplification is set by the
% Support setting (% Supp). PAV+ software continuously monitors the
patients instantaneous inspiratory flow and instantaneous lung volume, which
are indicators of the patients inspiratory effort. These signals, together with
ongoing estimates of the patients resistance and compliance, allow the software to instantaneously compute the necessary pressure at the patient wye to
assist the patients inspiratory muscles to the degree selected by the % Supp
setting. Higher inspiratory demand yields greater support from the ventilator.
PAV+ software reduces the risk of inadvertent entry of incompatible settings,
such as small predicted body weight (PBW) paired with a large airway.

D.2

Intended Use
PAV+ is intended for use in spontaneously breathing adult patients whose ventilator predicted body weight (PBW) setting is at least 25.0 kg (55 lb). Patients
must be intubated with either endotracheal (ET) or tracheostomy (Trach) tubes
of internal diameter (ID) 6.0 mm to 10.0 mm. Patients must have satisfactory
neural-ventilatory coupling, and stable, sustainable inspiratory drive.

*. Proportional Assist and PAV are registered trademarks of The University of Manitoba, Canada. Used under license.

D-1

PAV+ Appendix

D.3

Safety Symbol Definitions


This section contains safety information for users, who should always exercise
appropriate caution while using the ventilator.
TableD-1.Safety Symbol Definitions
Symbol

Definition
WARNING
Warnings alert users to potential serious outcomes (death, injury, or adverse
events) to the patient, user, or environment.
Caution
Cautions alert users to exercise appropriate care for safe and effective use of the
product.
Note
Notes provide additional guidelines or information.

D-2

WARNING:
The ventilator offers a variety of breath delivery options. Throughout the
patient's treatment, the clinician should carefully select the ventilation mode
and settings to use for that patient based on clinical judgment, the condition
and needs of the patient, and the benefits, limitations and characteristics of
the breath delivery options. As the patient's condition changes over time,
periodically assess the chosen modes and settings to determine whether or
not those are best for the patient's current needs.
WARNING:
PAV+ is not an available breath type in non-invasive ventilation (NIV). Do not
use non-invasive patient interfaces such as masks, nasal prongs, uncuffed ET
tubes, etc. as leaks associated with these interfaces may result in over-assist
and patient discomfort.
WARNING:
Breathing circuit and artificial airway must be free from leaks. Leaks may
result in ventilator over-assist and patient discomfort.

Operators Manual

Ventilator Settings/Guidance

D.4

WARNING:
Ensure high and low tidal volume alarm thresholds are set appropriately
because an overestimation of lung compliance could result in an undersupport condition resulting in the delivery of smaller than optimal tidal
volumes.

PAV+
WARNING:
Ensure that there are no significant leaks in the breathing circuit or around
the artificial airway cuff. Significant leaks can affect the performance of the
PAV+ option and the accuracy of resistance (R) and elastance (E) estimates.
WARNING:
Do not use silicone breathing circuits with the PAV+ option: the elastic
behavior of a silicone circuit at the beginning of exhalation can cause
pressure-flow oscillations that result in underestimates of patient resistance.

The act of inspiration requires the patients inspiratory muscles to develop a


pressure gradient between the mouth and the alveoli sufficient to draw in
breathing gas and inflate the lungs. Some of this pressure gradient is dissipated
as gas travels through the artificial airway and the patients conducting airways, and some of the pressure gradient is dissipated in the inflation of the
lungs and thorax. Each element of pressure dissipation is characterized by a
measurable property: the resistance of the artificial and patient airways, and
the compliance (or elastance) of the lung and thorax.
PAV+ software uses specific information, including resistance of the artificial
airway, resistance of the patients airways, lung-thorax compliance, instantaneous inspiratory flow and lung volume, and the % Supp setting to compute
the instantaneous pressure to be applied at the patient connection port
(patient wye). PAV+ software randomly estimates patient resistance and compliance approximately every four to ten breaths. Every five (5) ms, the software
estimates lung flow, based on an estimate of flow at the patient wye, and lung
volume, based on the integral of the value of estimated lung flow.
PAV+ begins to assist an inspiration when flow (generated by the patients
inspiratory muscles) appears at the patient wye. If the patient ceases inspiration, the assist also ceases. Once inspiratory flow begins, PAV+ software monitors instantaneous flow and volume every 5 ms and applies the pressure

Operators Manual

D-3

PAV+ Appendix

calculated to overcome a proportion (determined by the % Supp setting) of


the pressure losses dissipated across the resistances of the artificial and patient
airways and lung/thorax compliance.
Because the PAV+ algorithm does not know the patients mechanics when the
PAV+ breath type is selected, the software performs a startup routine to obtain
initial data. At startup, PAV+ software delivers four consecutive PAV+ breaths,
each of which includes an end-inspiratory pause maneuver that yields estimates of the patients resistance and compliance. The first breath, however, is
delivered using the predicted resistance for the artificial airway and conservative estimates for patient resistance and compliance, based on the patients
PBW.
Each of the next three PAV+ breaths averages stepwise decreased physiologic
values with the estimated resistance and compliance values from the previous
breath, weighting earlier estimates less with each successive breath, and yielding more reliable estimates for resistance and compliance. The fifth PAV+
breath (the first non-startup breath) is delivered using the final estimates with
the clinician-set % Supp setting. Once startup is complete, the PAV+ software
randomly applies a maneuver breath every four to ten breaths after the last
maneuver breath to re-estimate patient resistance and compliance. New values
are always averaged with former values.
The PAV+ option graphically displays estimates of patient lung pressure (intrinsic PEEP), patient compliance, patient resistance, total resistance, total work of
inspiration, patient work of inspiration, inspiratory elastic work (an indicator of
lung-thorax work), and inspiratory resistive work.
The % Supp setting ranges from a minimum of 5% (the ventilator performs
5% of the work of inspiration and the patient performs 95%) to a maximum
of 95% (the ventilator performs 95% of the work and the patient performs
5%), adjustable in 5% increments.
The PAV+ option also includes alarm limits, safety checks, and logic checks that
reject non-physiologic values for patient resistance and compliance as well as
inappropriate data.
Humidification type and volume can be adjusted after running SST, however
the ventilator makes assumptions when calculating resistance and compliance
if these changes are made without re-running SST. For optimal breath delivery,
run SST after changing humidification type and humidifier volume.

D-4

Operators Manual

Ventilator Settings/Guidance

D.4.1

Setting Up PAV+
To set up PAV+
1.

At the ventilator setup screen, enter the patients gender and height or the
patients PBW.

2.

Touch INVASIVE vent type.

3.

Touch SPONT mode.

4.

Touch PAV+ to select Spontaneous type.

5.

Touch the desired trigger type (P-TRIG, V-TRIG, or IE Sync (if the IE Sync option is
installed).

6.

Select tube type

7.

Select the tube ID. Initially, a default value is shown based on the PBW entered at
ventilator startup. If this ID is not correct for the airway in use, turn the knob to
adjust the ID setting.

Note:
If the operator selects an internal diameter that does not correspond to the PBW/
tube ID range pairs listed in the following table, touch the continue button to
override the tube ID setting. If attempts are made to choose a tube ID less than
6.0 mm or greater than 10 mm, a hard bound limit is reached, as PAV+ is not
intended for use with tubes smaller than 6.0 mm or larger than 10.0 mm. When
touching Dismiss, the setting remains at the last tube ID selected. Touch Accept
or Accept ALL to accept changes, or touch Cancel to cancel changes.
8.

Operators Manual

Continue setting up the ventilator as described in Chapter 4 of this manual.

Note:
If Leak Sync is currently enabled, it becomes disabled when PAV+ is selected.

D-5

PAV+ Appendix

FigureD-1.Ventilator Setup Screen

D.4.2

Note:
When the ventilator is used on the same patient previously ventilated using PAV+,
the GUI displays an attention icon and the tube type and tube ID previously used,
as a reminder to the clinician to review those settings during ventilator setup.

PBW and Tube ID


The ventilator uses soft bound and hard bound values for estimated tube
inside diameters based upon PBW. Soft bounds are ventilator settings that
have reached their recommended high or low limits. When adjusting the tube
size, if the inside diameter does not align with a valid predicted body weight,
a Continue button appears. Setting the ventilator beyond these soft bounds
requires the operator to acknowledge the prompt by touching Continue
button before continuing to adjust the tube size. The limit beyond which the
tube ID cannot be adjusted is called a hard bound, and the ventilator emits an
invalid entry tone when a hard bound is reached.

D-6

WARNING:
Ensure that the correct artificial airway ID size is entered. Because PAV+
amplifies flow, entering a smaller-than-actual airway ID causes the flow-

Operators Manual

Ventilator Settings/Guidance

based pressure assistance to over-support the patient and could lead to


transient over-assist at high values of % Supp. Conversely, entering a largerthan-actual ID results in under-support. PAV+ software monitors the settings
for the PBW and artificial airway. If the PBW and tube ID settings do not
correspond to the above PBW/tube ID range pairs, confirm or correct the
settings. Confirming or correcting the actual ID size minimizes the likelihood
that PAV+ will over-support or under-support.

To apply new settings for the artificial airway follow these steps
1.
Touch Vent Setup at the lower left of the GUI screen.
2.

Touch Tube Type and turn the knob to select Trach or ET to set the tube type.

3.

Touch tube ID and turn the knob to set the tube ID.

4.

Touch Accept or Accept ALL to apply the new settings, or Cancel to cancel.

To apply new humidifier settings


1.
Touch the More Settings tab.

D.4.3

2.

Touch the appropriate button for Humidification Type.

3.

For non-HME humidification types, touch Humidifier Volume, then turn the knob
to adjust the (empty) humidifier volume.

4.

Touch Accept ALL to apply the changes.

WARNING:
To ensure the accuracy of PAV+ breaths and spirometry measurements, run
SST following any change to the humidification type or humidification
volume settings. Ensure that the intended circuit is used with the SST.

Apnea Parameters Adjustment


After accepting the PAV+ settings, touch the Apnea Setup screen. Adjust the
Apnea parameters as required.

Operators Manual

D-7

PAV+ Appendix

D.4.4

Alarm Settings Adjustment


PAV+ includes the high inspired tidal volume 2VTI) and low exhaled spontaneous tidal volume alarm (4VTE SPONT) alarm limit settings. Reference PAV+
Alarms, p. D-10.

Note:
Because of the breathing variability that PAV+ allows, the 4VTE SPONT alarm, by
default, is turned OFF to minimize nuisance alarms. To monitor adequate ventilation,
use the 3VE TOT alarm condition instead.
To adjust alarm settings
1.
Touch the Alarm tab to view the current alarm settings.

D.4.5

2.

Touch the button for each alarm limit requiring a change.

3.

Turn the knob to adjust the value of the alarm limit. Proposed values are highlighted. You can change more than one alarm limit before applying the changes.

4.

Touch Accept or Accept All to apply the changes, or Cancel to cancel.

PAV+ Ventilator Settings


Reference the table Ventilator Settings Range and Resolution, in Chapter 11 of
this manual for a summary of PAV+ ventilator settings for the following parameters:

D.4.6

% Supp

Expiratory sensitivity (ESENS)

Tube type

Tube ID

Trigger type

PAV+ Alarm Settings


Reference the table Alarm Settings Range and Resolution, in Chapter 11 of this
manual for a summary of the following alarm settings available when PAV+ is
active:

D-8

Operators Manual

Ventilator Settings/Guidance

D.4.7

High inspired tidal volume limit (2VTI)

Low exhaled spontaneous tidal volume (4VTE SPONT)

Monitored Data
Reference Patient Data Range and Resolution, in Chapter 11 of this manual for
the following monitored data associated with PAV+:

PAV-based lung compliance (CPAV)

PAV-based lung elastance (EPAV)

PAV-based lung resistance (RPAV)

PAV-based total airway resistance (RTOT)

Inspired tidal volume (VTI)

Reference the table below for monitored data absolute limits.


TableD-2.Absolute limits for PAV+ Monitored Data

Operators Manual

PBW (kg)

RPAV (cmH2O/L/s)

CPAV (mL/cmH2O

EPAV (cmH2O/L

25

0 to 50

2.5 to 29

34 to 400

35

0 to 44

3.5 to 41

24 to 286

45

0 to 31

4.5 to 52

19 to 222

55

0 to24

5.5 to 64

16 to 182

65

0 to 20

6.4 to 75

13 to 156

75

0 to 18

7.4 to 87

11 to 135

85

0 to 17

8.4 to 98

10 to 119

95

0 to 16

9.4 to 110

9.1 to 106

105

0 to 15

10 to 121

8.3 to 100

115

0 to 15

11 to 133

7.5 to 91

125

0 to 14

12 to 144

6.9 to 83

135

0 to 14

13to 156

6.4 to 77

145

0 to 14

14 to 167

6.0 to 71

150

0 to 14

15 to 173

5.8 to 67

D-9

PAV+ Appendix

D.4.8

PAV+ Alarms
Reference Non-technical Alarm Summary in Chapter 6 of this manual for a
summary of the following alarms associated with PAV+:

D.5

High circuit pressure (1PPEAK)

High ventilator pressure(1PVENT)

PAV STARTUP TOO LONG

PAV R&C NOT ASSESSED

1VTI

Ventilator Settings/Guidance
WARNING:
For optimal performance of PAV+, it is important to select the humidification
type, tube type, and tube size that match those in use on the patient.

The instantaneous pressure generated at the patient wye during inspiration is


a function of the patient effort, % Supp setting, tube type and size, patient
resistance and elastance, and the instantaneously measured gas flow and lung
volume. Set 2PPEAK to a safe circuit pressure, above which truncation and
alarm annunciation are appropriate.

D-10

Note:
PAV+ has a built-in high pressure compensation (1PCOMP) limit that is determined by
the2PPEAK setting minus 5 cmH2O or 35 cmH2O, whichever is less. If the inspiratory
pressure at the patient wye (PI wye) reaches the 1PCOMP limit, the inspiration is
truncated, and the ventilator transitions to exhalation. Reference p. D-21 for more
details regarding 1PCOMP and 1PPEAK.

Operators Manual

Ventilator Settings/Guidance

D.5.1

Specified performance

Performance using PAV+ is 0.5 Joules/liter (J/L), compared to measured,


work during inspiration at the 75% support (% Supp) level. Work is computed
over the entire inspiratory interval. In ventilation terms, work (W) is expressed
as:

k P V dt
W = -------------------------------------- i dt
V

ith sample interval (5 ms)

Flow [L/s]

Work [J/L]

conversion constant (0.098) [J/cmH2O x L)

Synchronous and combined pressures developed by the ventilator and


by the patient (PMUS), [cmH2O]

D.5.2

Graphics Displays in PAV+


When PAV+ is active (the mode is SPONT and the spontaneous breath type is
PAV+), a work of breathing (WOB) graphic is automatically displayed (Reference Graphics displays in PAV+, p. D-14) which shows:

Operators Manual

estimates of work of breathing relative to normal, subnormal, and above-normal


values, including:

the estimated work of breathing in Joules/L) during inspiration (WOBPT) and

the estimated total work of breathing (in Joules/L) of the patient and ventilator during inspiration (WOBTOT)

D-11

PAV+ Appendix

an indicator showing the proportion of patient inspiratory work to overcome the


elastance (E) of the lung-thorax and the combined resistance (R) of the artificial
airway and the patient.

Additional information in the graphics screen includes

a shadow trace of the estimated lung pressure, shown as a solid area superimposed on the circuit pressure waveform.

PAV-based patient data estimates, including patient resistance (RPAV), lung compliance (CPAV), and intrinsic PEEP (PEEPI PAV).

Note:
Graphic displays of lung pressure and patient work of breathing are not actual
measurements, and are derived from equations using filtered estimates of pressure
and flow.

The WOB graphic is only available when SPONT mode and the PAV+ breath
type are selected. The shadow trace can be enabled or disabled when selecting
the graphic display, or after a display is paused.
The act of pausing does not affect the WOB graphic, but does store the
shadow trace. Once paused, the operator can enable or disable the shadow
trace, then view the paused waveform again with or without the shadow
trace.
D.5.3

WOB Terms and Definitions


The following table provides a definition and description of each of the Work
of Breathing (WOB) terms.

D-12

Operators Manual

Ventilator Settings/Guidance

TableD-3.PAV+ Work of Breathing terms


WOB term

Definition

Description

WOBTOT

Total Work of Inspiration

With the PAV+ breath type


active, the patient and the ventilator always share the in the
work of breathing. The percent
WOBTOT performed by the ventilator always equals the %
Supp setting and the percent
WOBTOT performed by the
patient always equals (100
minus the % Supp setting).
WOBTOT is the sum of the work
to move the breathing gas
through the artificial airway
and the patient's own airways
plus the work to inflate the
patient's elastic lung-thorax.

WOBPT

Patient Work of Breathing

That part of WOBTOT performed by the patient.

WOBPT ELASTIC

Inspiratory Elastic Work

That part of WOBPT attributed


to inflating the patients elastic
lung-thorax.

WOBPT RESISTIVE

Inspiratory Resistive Work

That part of the WOBPT


attributed to moving breathing
gas through resistive elements
in the gas path.

Operators Manual

D-13

PAV+ Appendix

FigureD-2.Graphics displays in PAV+

D.5.4

Total work of breathing (WOBTOT)

Patients work of breathing (WOBPT)

Shadow trace

Technical Description
When PAV+ is selected, the ventilator acts as an inspiratory amplifier, proportionally assisting the pressure generating capability of the inspiratory muscles
(PMUS).
Pressure Gradient Equation of Motion
During spontaneous breathing, PMUS generates a pressure gradient that drives
breathing gas through the artificial airway and the patients airways and into
the elastic lung-thorax, and is described by the equation of motion:

D-14

Operators Manual

Ventilator Settings/Guidance

EQUATION 1

P MUS = V L R + V L E LUNG THORAX

PMUS

Pressure generating capability of


patients inspiratory muscles

Resistance elements (artificial plus patient airways)

VL

Flow through the resistance elements and into the lungs

ELUNG-THORAX

Elastance of the lung and thorax (1/CLUNGTHORAX)

VL

Insufflation volume of the lung

Operators Manual

D-15

PAV+ Appendix

Estimates of Patient Resistance and Elastance


If the PAV+ software estimates of patient resistance and elastance (RPAV and
(EPAV) remain stable, this equation could be rewritten as:
EQUATION 2

i
i
i
i
i
P MUS = V L R airway + V L K 1 + V L K 2

Instantaneous value of pressure, flow,


or airway resistance, Riairway being a
function of flow

K2

EPAV

K1

RPAV

PiMUS could then be estimated at every control period if ViL, Riairway, and ViL
were also known.
Valid Individual Pressure Measurements
Throughout any inspiration, the individual pressure elements that make up
PMUS can be expressed as:
EQUATION 3

FLOW

p MUS = P ARTIFICIAL

VOLUME

+ P PATIENT + P PATIENT

PMUS

Pressure generating capability of


patients inspiratory muscles

PFLOWPATIENT

Flow based pressure drop across the


patient

PFLOWARTIFI-

Flow based pressure drop across


the artificial airway

PVOLUMEPATIENT

Volume based pressure to overcome


the lung-thorax elastance

CIAL AIRWAY

D-16

FLOW

AIRWAY

Operators Manual

Ventilator Settings/Guidance

Equations 2 and 3 provide the structure to explain how PAV+ operates. The
clinician enters the type and size of artificial airway in use, and the software
uses this information to estimate the resistance of the artificial airway at any
lung flow.
Applying a special pause maneuver at the end of selected inspirations provides
the information the software needs to estimate patient resistance (RPAV) and
compliance (CPAV, which is converted to elastance, EPAV). Immediately following the end of the pause event, software captures simultaneous values for
PLUNG, Pwye, and VE which yield an estimate for RTOT at the estimated flow.
All raw data are subjected to logic checks, and the estimates of RPAV and CPAV
are further subjected to physiologic checks. The estimates of RPAV and CPAV are
discarded if any of the logic or physiologic checks fail. If CPAV is rejected, RPAV
is also rejected.
Valid estimates of RPAV and CPAV are required for breath delivery, and are constantly updated by averaging new values with previous values. This averaging
process smooths data and avoids abrupt changes to breath delivery. If new
values for RPAV and CPAV are rejected, the previous values remain active until
valid new values are obtained. PAV+ software monitors the update process
and generates an escalating alarm condition if the old values do not refresh.
Maneuver Breaths and % Supp
During PAV+, maneuver breaths are randomly performed every four to ten
breaths after the last maneuver breath. A maneuver breath is a normal PAV+
inspiration with a pause at end inspiration. Because muscle activity is delayed
for at least 300 ms following the end of neural inspiration, the patients respiratory control center does not detect the pause. With this approach, maneuver
breaths are delivered randomly so that their occurrence is neither consciously
recognized nor predictable.
A PAV+ breath begins, after the recognition of a trigger signal, with flow
detection at the patient wye. The sample and control cycle of the ventilator
(the value of i in Equation 2) is frequent enough to yield essentially constant
tracking of patient inspiration. At every ith interval, the software identifies
instantaneous lung flow (ViL), which is impeded by the resistances of the artificial airway and patient airways) and integrates this flow to yield an estimate
of instantaneous lung volume, (ViL), which is impeded by the elastic recoil of
the lung and thorax).

Operators Manual

D-17

PAV+ Appendix

Using the values for instantaneous lung flow and lung volume, PAV+ software
calculates each of the pressure elements in Equation 2, which gives the value
of PMUS at each ith interval.
At this point, Equation 2 and the subsequent analysis identifies that an appropriate patient, supported by PAV+ and with an active PMUS (an absolute
requirement) will, within a few breaths, enable the algorithm to obtain reasonable estimates of RPAV and EPAV. Once these physiologic data are captured
(and over a relatively brief time as they are improved and stabilized), the PAV+
algorithm mirrors the patient's respiratory mechanics, which then allows the
ventilator to harmoniously amplify PMUS. The key point to recognize is that
patient's continuous breathing effort drives the PAV+ support no effort,
no support.
The % Supp setting specifies the amount of resistance- and elastic-based pressure to be applied at each ith interval at the patient wye.
By taking all of the above information into consideration, EQUATION 2 can be
rewritten to include the % Supp setting recognizing that ViL and ViL are driven
by the patient, not by the ventilator. (It is important to note that the ventilator
is not amplifying its own flow only the flow generated by PMUS.)

EQUATION 4

i
i
i
i
i
P wye = S V L R airway + S V L K 1 + S V L K 2

Piwye

D-18

Pressure generated by the ventilator in


response to the instantaneous values of
lung flow and lung volume at the wye.
This value is the sum of the three individual pressure elements (in parentheses) in Equation 4

% Supp setting/100 (ranges from 0.05


to 0.95

Operators Manual

Ventilator Settings/Guidance

Resulting Pressure Gradient


The pressure gradient driving breathing gas into the patients lungs is given by
the sum of Piwye and the patients inspiratory effort, therefore:
EQUATION 5

PiMUS

Instantaneous pressure generating


capability of patients inspiratory
muscles

P GRADIENT = P wye + P MUS

PiGRADIENT

Instantaneous pressure gradient

piwye

Pressure generated by the ventilator in


response to the instantaneous values of
lung flow and lung volume at the wye

D.5.5

Protection Against Hazard


PAV+ software is designed to reduce the risk that hyperinflation may occur.
The potential for hyperinflation could arise if the software were to overestimate actual patient resistance or underestimate actual patient lung-thorax
compliance (that is, to overestimate actual elastance). If the software cannot
generate valid estimates of RPAV and CPAV, PAV+ cannot start. If, after startup,
the values of RPAV and CPAV cannot be updated with valid new values, the previous values become less reliable.
The stability of PAV+ is primarily determined by the relationship between the
true lung elastance [EL (true)] and the true lung volume [VL(true)]. Although Pi
wye (resistive) also plays a part, the following discussion focuses on the elastic
component.
At all lung volumes, the true state of the lung and thorax is expressed by:

Operators Manual

D-19

PAV+ Appendix

P L recoil = V L true E L true

PiL recoil

True lung recoil pressure

EL (true)

ViL (true)

True instantaneous volume of the lung

True lung elastance

Over-inflation will not occur as long as Piwye (elastic) < PiL recoil, which is equivalent to the inequality:
S[ViL (estimated) x K2] < ViL (true) x EL(true)

where:
K2 = EPAV1
1. see equations 2 and 4

As long as EPAV (estimated) = EPAV (true) and ViL (estimated) = ViL (true) then
Pirecoil> Piwye even at high values of % Supp (i.e. between 85% and 95%).
This means that if the pressure applied to the lung-thorax is never greater than
EL (true) x VL, lung volume will collapse if wye flow vanishes. As long as EPAV
(estimated) EL (true), ViL (estimated) ViL (true), and RPAV (estimated)
RL(true), PMUS is the modulator of Piwye.
Hyperinflation could occur if the estimated EPAV were greater than the true
value of EL. At a high % Supp setting, Piwye (elastic) could exceed PiL recoil,
causing a self-generating flow at the patient wye, which in turn would cause
a self-generating inflation of the lungs. This is part of the reason that the
% Supp setting is limited to 95%.
Likewise, if the estimated RPAV were to exceed the true value of RL at a high
% Supp setting, Piwye (resistive) could exceed the value necessary to compensate for pressure dissipation across the artificial and patient airways, resulting
in early hyperinflation of the lungs. As flow declines after the first third of inspiration, however, the hyperinflating effect would most likely disappear.

D-20

Operators Manual

Ventilator Settings/Guidance

PAV+ software includes these strategies to minimize the possibility of hyperinflation of the lungs:
1.

The maximum % Supp setting is limited to 95%.

2.

The raw data for RPAV and CPAV are checked for graph/math logic, and estimated
mechanics values are checked against PBW-based physiologic boundaries. These
checks reduce the possibility of overestimating patient resistance or underestimating patient compliance, which could lead to potential over-inflation.

3.

The high inspiratory tidal volume limit (2VTI) places an absolute limit on the integral of lung flow (including leak flow), which equals lung volume. If the value of
VTI reaches this limit, the ventilator truncates inspiration and immediately transitions to exhalation.

4.

The 2VTI setting places an upper limit on the value of the PVOLUMEPATIENT component of Piwye (see Equations 3 and 4). At the beginning of each new inspiration,
PAV+ software calculates a value for PVOLUMEPATIENT as follows:

P*wye (elastic threshold limit) = 0.75 x (VTI x EPAV)


where P*wye is the unique value for the elastic threshold limit of Piwye that will
cause the lung volume to expand to 75% of 2VTI. When Piwye (elastic) = P*wye
(elastic threshold limit), the software stops increasing Piwye (elastic). This means
that any further increase in lung volume must be accomplished by the patient,
which tends to hasten the conclusion of inspiratory effort and avoid truncation
due to lung volume reaching the 2VTI limit.
5.

The high inspiratory pressure limit (2PPEAK) applies to all breaths, and is used by
PAV+ software to detect the high compensation pressure condition (1PCOMP):

1PCOMP = 2PPEAK - 5 cmH2O or 35 cmH2O, whichever is less


If the user-adjustable 2PPEAK limit is reached, the ventilator truncates inspiration
and immediately transitions to exhalation. If Piwye (the targeted wye pressure calculated in Equation 4) equals the 1PCOMP for 500 ms, the inspiration is truncated
and exhalation begins. Further, when Piwye = 1PCOMP, Piwye is limited to 1PCOMP.

Operators Manual

D-21

PAV+ Appendix

Although this freezes the value of Piwye, patient activity such as coughing could
drive Piwye to 2PPEAK, causing inspiration to end.
The rapid rise of Piwye to the 1PCOMP limit would likely occur in the first third of
inspiration, and only if RPAV were overestimated and % Supp were set above
85%. The 1PCOMP condition guards against over-inflation due to overestimation
of RPAV.
6.

The % Supp setting ranges from 5 to 95% in 5% increments. Reducing the level
of support decreases the possibility of over-inflation. A significant decrease could
produce a sensation of inadequate support, and the patient would absorb the
additional work of inspiration or require an increase in the level of support.
A significant increase could cause a surge in the ventilator generated value for
Pwye, which in turn could cause Piwye to reach 2PCOMP and lead to temporary
patient-ventilator disharmony. To minimize this possibility, PAV+ software limits
the actual increase in support to increments of 10% every other breath until the
new setting is reached.

7.

Spirometry remains active during PAV+ operation. 2VTI can be set high enough to
allow spontaneous sigh breaths, while 4VE TOT and 2VE TOT remain active to reveal
changes in minute ventilation.
Because PAV+ cannot operate without valid estimates of RPAV and CPAV, and
because those values are unknown when PAV+ starts, a startup routine obtains
these values during four maneuver breaths that include an end inspiratory pause
that provides raw data for RPAV and CPAV, and both estimated values must be
valid. If either value is invalid during any of the four startup breaths, the software
schedules a substitute maneuver breath at the next breath. Reference PAV+, p.
D-3.
A low-priority alarm becomes active if a 45-second interval elapses without valid
estimates for RPAV and CPAV. If the condition persists for 90 seconds, the alarm
escalates to medium-priority. If the condition persists for 120 seconds, the alarm
escalates to high priority. The 3VE TOT and1fTOT alarms are also associated with this
condition.
Similarly, if RPAV and CPAV cannot be updated with valid values after a successful
PAV+ startup, a low-priority alarm is activated if the condition persists for 15 minutes. If the values still cannot be updated with valid values after 30 minutes, the
alarm escalates to medium priority.

D-22

Operators Manual

Ventilator Settings/Guidance

If PAV+ estimates a high lung resistance following a sharp spike in the expiratory
flow waveform, then a PBW-based resistance value is used. Reference the waveform and table below.

8.

FigureD-3. Use of Default Lung Resistance

Flow (V)

High peak expiratory flow

Expiration

Exhalation with normal return to zero


flow

Inspiration

Normal peak expiratory flow

Exhalation with slow, restricted return


to zero flow

TableD-4.Default PBW-based Resistance Values


PBW
(kg)

Resistance
(cmH2O/L/s)

PBW
(kg)

Resistance
(cmH2O/L/s)

PBW
(kg)

Resistance
(cmH2O/L/s)

PBW (kg)

25

18.1

43

13.5

61

11.3

79

10.1

26

17.7

44

13.3

62

11.2

80

10.1

27

17.4

45

13.2

63

11.1

81 to 150

28

17.1

46

13.0

64

11.0

29

16.8

47

12.9

65

10.9

30

16.5

48

12.7

66

10.9

31

16.2

49

12.6

67

10.8

32

15.9

50

12.4

68

10.7

33

15.7

51

12.3

69

10.7

34

15.4

52

12.2

70

10.6

Operators Manual

Resistance
(cmH2O/L/s)

10

D-23

PAV+ Appendix

TableD-4.Default PBW-based Resistance Values (Continued)

D-24

PBW
(kg)

Resistance
(cmH2O/L/s)

PBW
(kg)

Resistance
(cmH2O/L/s)

PBW
(kg)

Resistance
(cmH2O/L/s)

35

15.2

53

12.1

71

10.5

36

14.9

54

12.0

72

10.5

37

14.7

55

11.8

73

10.4

38

14.5

56

11.7

74

10.4

39

14.3

57

11.6

75

10.3

40

14.1

58

11.5

76

10.3

41

13.9

59

11.4

77

10.2

42

13.7

60

11.3

78

10.2

PBW (kg)

Resistance
(cmH2O/L/s)

Operators Manual

E NeoMode 2.0 Appendix

E.1

Overview
This appendix describes how to use NeoMode 2.0 software on the Puritan Bennett 980 Neonatal Ventilator. NeoMode 2.0 enables the use of the ventilator with neonatal patients and is included with all Puritan Bennett 980
Neonatal Ventilators, and Puritan Bennett 980 Universal Ventilators. In
order for a Puritan Bennett 980 Pediatric-Adult Ventilator to be used with
Neonatal patients, the NeoMode 2.0 software option must be installed.

E.2

Intended Use
NeoMode 2.0 is intended to provide respiratory support to neonatal patients
with predicted body weights as low as 0.3 kg (0.66 lb). It is intended to cover
a wide variety of clinical patient conditions in hospitals and hospital-type facilities, and may be used during intra-hospital and intra-hospital-type facility
transport. It supports delivered tidal volumes as low as two (2) mL.

E.3

Safety Symbol Definitions


This section contains safety information for users, who should always exercise
appropriate caution while using the ventilator.

E-1

NeoMode 2.0 Appendix

TableE-1.Safety Symbol Definitions


Symbol

Definition
WARNING
Warnings alert users to potential serious outcomes (death, injury, or adverse
events) to the patient, user, or environment.
Caution
Cautions alert users to exercise appropriate care for safe and effective use of the
product.
Note
Notes provide additional guidelines or information.

E.4

E-2

Safety Information
WARNING:
The Puritan Bennett 980 Series Ventilator contains phthalates. When used
as indicated, very limited exposure to trace amounts of phthalates may occur.
There is no clear clinical evidence that this degree of exposure increases
clinical risk. However, in order to minimize risk of phthalate exposure in
children and nursing or pregnant women, this product should only be used as
directed.
WARNING:
The ventilator offers a variety of breath delivery options. Throughout the
patient's treatment, the clinician should carefully select the ventilation mode
and settings to use for that patient based on clinical judgment, the condition
and needs of the patient, and the benefits, limitations and characteristics of
the breath delivery options. As the patient's condition changes over time,
periodically assess the chosen modes and settings to determine whether or
not those are best for the patient's current needs.
WARNING:
Neonatal patients are at risk for hypercarbia or hypoxemia during
3VE TOT alarm conditions.

Operators Manual

Ventilation Features

E.5

WARNING:
Disabling the low exhaled minute volume (3VE TOT) alarm increases the
patients risk of Hypercarbia or Hypoxemia.
WARNING:
When Using NIV, the patients exhaled tidal volume (VTE) could differ from
the ventilators monitored patient data value for VTE due to leaks around the
interface. To avoid this, ensure Leak Sync is installed. When NIV is selected,
Leak Sync is automatically enabled.

Description
The ventilator determines values for operational variables and allowable settings based on breathing circuit type and predicted body weight (PBW). The
PBW range for neonates is 0.3 kg to 7.0 kg (0.66 lb to 15 lb). Software controls
prevent inadvertent mismatching of patient size and breathing circuit type. A
neonatal breathing circuit connects to a neonatal expiratory filter which must
be used with the neonatal adapter door assembly.

E.6

Note:
To enable NeoMode 2.0, select the neonatal breathing circuit type in Short Self Test
(SST). Breathing circuit type can only be changed during SST. NeoMode 2.0 software
restricts the circuit type selection to neonatal.

Neonatal Door and Filter Installation


WARNING:
Removing the expiratory filter while the patient is connected to the ventilator
can cause a loss of circuit pressure, ventilator autotriggering, or direct contact
with liquid.
Caution:
Do not pull on door while expiratory filter latch is closed, as damage to the
ventilator can result.

Operators Manual

E-3

NeoMode 2.0 Appendix

Note:
Reference the Inspiratory filter and expiratory filter instructions for use (IFU) for
information on filtration efficiency and filter resistance.
To install the neonatal adapter door
1.
Remove expiratory limb of patient circuit from expiratory filter.
2.

Lift expiratory filter latch. Reference Installing Neonatal Filter and Door, p. E-4.

3.

Remove existing expiratory filter door. by lifting it off of pivot pins.

4.

Fit neonatal adapter door onto pivot pins.

To install neonatal expiratory filter assembly


1.
With the door still open, push the neonatal filter straight up into the adapter.
2.

Close the door.

3.

Lower the expiratory filter latch.

4.

Re-attach expiratory limb of patient circuit to filter.


FigureE-1.Installing Neonatal Filter and Door

E-4

Neonatal expiratory filter

Expiratory filter latch

Neonatal adapter door

Filter door pivot pin

Operators Manual

Ventilation Features

E.7

WARNING:

To ensure all breathing circuit connections are leak tight, perform a circuit
leak test by running SST every time the filter is installed. The circuit leak
test can be performed as an individual test from the SST startup screen,
but after the leak test is performed, the entire suite of SST tests must be
performed without a failure before ventilation can begin.

Empty the condensate vial before fluid reaches the maximum fill line. Condensate vial overflow can enter the filter or the breathing circuit, and can
cause increased expiratory flow resistance. Change the filter if it appears
to be saturated.

The neonatal expiratory filter and condensate vial is a single unit and is
for single-patient use, only. Do not attempt to sterilize the filter assembly.

WARNING:
Adding accessories to the ventilator can change the pressure gradient across
the ventilator breathing system (VBS) and affect ventilator performance.
Ensure any changes to the recommended ventilator circuit configurations do
not exceed the specified values for inspiratory and expiratory pressure drop.
Reference the table Technical Specifications on p. 11-3 in this manual. If
adding accessories to the patient circuit, always run SST to establish circuit
compliance and resistance prior to ventilating the patient.
Note:
If the ventilator has not reached operating temperature from recent usage, allow it to
warm up for at least 15 minutes before running SST to ensure accurate testing.
Note:
Check the inspiratory and expiratory limbs of the breathing circuit and in-line water
traps regularly for water buildup. Under certain conditions, they can fill quickly. Empty
and clean the in-line water traps as necessary. Reference the manufacturers
instructions for additional information.

How to Empty the Condensate Vial


The condensate vial may accumulate liquid, especially if a non-heated wire
patient circuit is in use.

Operators Manual

E-5

NeoMode 2.0 Appendix

WARNING:
To avoid liquid entering the ventilator, empty the condensate vial before the
liquid level reaches the maximum fill line.

The condensate vial assembly is integrated with the neonatal expiratory filter
and does not contain a drain port. Empty the condensate vial when fluid
reaches the maximum fill line.
To empty the condensate vial
1.
While holding the expiratory filter, twist the condensate vial clockwise approximately one quarter turn to remove it.

E.8

2.

Remove by carefully lowering the vial all the way down to the base of the exhalation compartment and then slide the vial out.

3.

Quickly empty the condensate vial.

4.

Replace the condensate vial by carefully sliding the vial into position, lifting it
upward to the filter assembly, and turning counterclockwise until it reaches the
stop.

Note:
Condensate vial removal may cause the loss of system pressure and a disconnect
alarm.

Connecting the Breathing Circuit


WARNING:
Use one of the ventilator breathing circuits listed, Reference Recommended
Breathing Circuits, p. E-8 or their equivalent. This ensures that maximum
pressure/flow values specified by EN794-1 are not exceeded. Using a circuit
with a higher resistance does not prevent ventilation, but can cause an SST
fault or compromise the patients ability to breathe through the circuit.

Reference How to Connect the Breathing Circuit, p. E-7 to connect the breathing circuit.

E-6

Operators Manual

Ventilation Features

FigureE-2.How to Connect the Breathing Circuit

Humidifier

From Patient port

Patient circuit inspiratory limb

Neonatal expiratory filter (installed in


adapter door)

Patient circuit wye

To Patient port

Patient circuit expiratory limb

Inspiratory filter

Condensate vial

Operators Manual

E-7

NeoMode 2.0 Appendix

TableE-2.Recommended Breathing Circuits


Patient Circuit

E.9

Part number

Ventilator breathing circuit, Neonatal disposable,


DAR (not available in USA)

3078447

Ventilator breathing circuit, Neonatal, disposable, Hudson RCI/Teleflex

780-06

Ventilator breathing circuit, Neonatal, disposable, Fisher & Paykel

RT235

Ventilation Features
Ventilation using NeoMode 2.0 is performed exactly as described in Chapter
4 of this manual. Reference Ventilator Operation, p. 4-7. If using a Puritan Bennett 980 Neonatal Ventilator, NeoMode 2.0 is already in use and a neonatal
patient circuit is the only choice available when performing SST. If using a
Puritan Bennett 980 Pediatric-Adult Ventilator or Puritan Bennett 980Universal Ventilator, the NeoMode 2.0 software option must be installed and SST
must be run using a neonatal patient circuit.

E.9.1

WARNING:
Always run SST prior to patient ventilation, ensuring that all accessories used
during ventilation are in the ventilator breathing system when SST is run. This
ensures correct calculation of compliance and resistance. Reference SST (Short
Self Test), p. 3-46 for more information.

Predicted Body Weight (PBW) vs. Patient Length


Reference Predicted Body Weight (PBW) Calculation in Chapter 4 of this
manual for tables of PBW values associated with patient length in centimeters
and inches, respectively.

E.9.2

Elevate O2
In NeoMode 2.0, the elevate O2 control works exactly as described in
Chapter 3 of this manual, except when the ventilator is in stand-by or circuit
disconnect states. If the elevate O2 function is used during these conditions,

E-8

Operators Manual

Ventilation Features

the value chosen for elevate O2 applies to the currently delivered oxygen concentration (which is 40% O2 in these cases) and not the set oxygen concentration.
E.9.3

CPAP Mode
When using NeoMode 2.0 and ventilating with non-invasive ventilation (NIV),
a separate CPAP mode allows spontaneous breathing with a desired PEEP
level. In order to limit inadvertent alarms associated with the absence of
returned volumes in nasal CPAP breathing, CPAP does not make available
exhaled minute volume and exhaled tidal volume alarm settings. Since some
neonates dont trigger breaths, the default apnea interval, TA, is set to OFF.
Also, some settings changes will initiate a PEEP restoration breath before
phasing in those changes.

Note:
In CPAP, Apnea Time, TA can be adjusted, if desired. It merely defaults to OFF to avoid
inadvertent alarms. The message APNEA DETECTION DISABLED is displayed at

the bottom of the GUI screen. The attention icons are also displayed.
To set the ventilator for CPAP
1.
Select New Patient from the ventilators startup screen or touch Vent Setup.
2.

Touch PBW and turn the knob to set the PBW.

3.

Select NIV as the Vent type.

4.

Touch CPAP.

5.

Touch each ventilator setting and turn the knob to select appropriate ventilator
settings. When finished, touch Accept or Accept ALL.

6.

Complete the setup by setting the apnea parameters and alarm limits from their
respective tabs.

Operators Manual

E-9

NeoMode 2.0 Appendix

FigureE-3.CPAP Setup Screen

E.9.4

Entering CPAP From Other Ventilation Modes


Entering CPAP mode from other ventilation modes or vent types requires using
the NIV Vent type.The section, Non-invasive Ventilation (NIV) in Chapter 4 of
this manual explains how the ventilator transitions from INVASIVE TO NIV vent
type.
To enter CPAP mode for an existing patient
1.
Touch Vent Setup at the lower left of the GUI screen.

E-10

2.

Touch NIV Vent type. CPAP is only allowed during NIV.

3.

Touch CPAP. The ventilator enters CPAP mode.

4.

Set an apnea interval, TA, if appropriate, as it defaults to OFF in CPAP.

Note:
Exhaled minute volume (VE TOT), exhaled tidal volume (VTE SPONT), and inspired tidal
volume (VTI) alarms are disabled upon entry into CPAP.

Operators Manual

Ventilation Features

E.9.5

Exiting CPAP Mode


When changing the mode from CPAP to any other mode, several transition
rules take effect:

E.9.6

The VE TOT, VTE MAND, VTE SPONT, and VTI alarms are set to their respective new
patient defaults.

When transitioning from apnea interval set to OFF to an apnea interval of time
(TA), the new setting is phased in immediately.

The VE TOT, VTE MAND, VTE SPONT, and VTI alarm sliders appear in the alarm settings
screen according to their applicability to the selected mode.

Compliance Compensation
Reference Compliance Compensation in Volume-based Breaths in Chapter 10
of this manual for a complete discussion of compliance compensation. Compliance compensation in NeoMode 2.0 is implemented exactly as described in
the aforementioned reference.

E.9.7

Note:
If the patients compliance decreases beyond the limits of compliance compensation,
the ventilator relies on the 2PPEAK alarm setting to truncate the breath and switch to
exhalation.

Settings, Alarms, and Monitored Patient Data


WARNING:
Monitor the patient closely if alarms are disabled: there are no audible or
visual annunciations for out-of-range conditions when volume, pressure, or
apnea alarms are disabled (turned OFF).

Reference the tables for Ventilator Settings, Alarm Settings, and Patient Data
in Chapter 11 of this manual for the ranges for each ventilator setting or data
value. Most settings, however, are also limited by other settings or conditions
(for example, a low alarm limit is always limited by the corresponding high
alarm limit). Review the prompt area when making settings changes.

Operators Manual

E-11

NeoMode 2.0 Appendix

Volume Accuracy
Volume accuracy testing in VC+ was conducted to demonstrate performance
of delivered and monitored parameters. The following tables provide a
summary of the actual results obtained within the range of two (2) mL to 25
mL collected during test execution.
The first column (Set Tidal Volume) represents the desired volume setting in
milliliters (mL). The second column represents the total number of test points
for test cases executed at that specific setting. The third column is the mean
(Mean Value) of the ventilators and test cases executed for the setting listed.
The fourth column represents the standard deviation of measurements taken
for the ventilators and test cases executed at the setting listed.
The measurements were taken using instrumentation located at the patientconnection port. Accessories such as filters and humidifiers were in the circuit
during the test. Values were BTPS and compliance compensated. A sample size
of five (5) ventilators was used to conduct the testing. Testing was conducted
at 22C 5C.
TableE-3.Delivered Volume Accuracy
Set Tidal Volume
(mL)

Number of Test
Points

Mean Value

STD

150

2.067

0.198

270

4.853

0.324

15

240

15.108

0.383

25

270

24.608

0.607

TableE-4.Monitored Inspired Volume (VTI) Accuracy

E-12

Set Tidal Volume


(mL)

Number of Test
Points

Mean Value

STD

150

2.055

0.192

270

4.872

0.346

15

240

15.235

0.379

25

265

24.633

0.566

Operators Manual

Ventilation Features

TableE-5.Monitored Exhaled Tidal Volume (VTE) Accuracy

Operators Manual

Set Tidal Volume


(mL)

Number of Test
Points

Mean Value

STD

150

2.212

0.274

270

5.892

0.607

15

240

16.145

0.851

25

265

25.492

0.819

E-13

NeoMode 2.0 Appendix

Page Left Intentionally Blank

E-14

Operators Manual

F Proximal Flow Option Appendix

F.1

Overview
This appendix describes the operation of the Proximal Flow Option for the
Puritan Bennett 980 Series Ventilator. The Proximal Flow Option is solely
used for monitoring flows, pressures, and tidal volumes and does not control
these parameters in any way.
The Proximal Flow Sensor is designed to measure the lower flows, pressures
and tidal volumes at the patient wye typically associated with invasively ventilated neonatal patients.
For general parameter and general ventilator setup information, reference
Chapter 4 in this manual.

F.2

Intended Use
The Proximal Flow Option is used for measuring flows, pressures, and tidal
volumes of invasively ventilated neonatal patients with predicted body weights
(PBW) of 0.3 kg (0.66 lb) to 7.0 kg (15.4 lb) using ET tube sizes from 2.5 mm
to 4.0 mm. The NeoMode 2.0 software option must also be installed on the
ventilator.

F.3

Safety Symbol Definitions


This section contains safety information for users who should always exercise
appropriate caution while using the ventilator.

F-1

Proximal Flow Option Appendix

TableF-1.Safety Symbol Definitions


Symbol

Definition
WARNING
Warnings alert users to potential serious outcomes (death, injury, or adverse
events) to the patient, user, or environment.
Caution
Cautions alert users to exercise appropriate care for safe and effective use of the
product.
Note
Notes provide additional guidelines or information.

F.4

Software/Hardware Requirements
The Proximal Flow Option requires installation of the NeoMode 2.0 software
option or a Puritan Bennett 980 Neonatal Ventilator must be used. Details
regarding NeoMode 2.0 can be found in the NeoMode 2.0 Appendix .

F.5

F-2

Safety Information
WARNING:
The Puritan Bennett 980 Series Ventilator contains phthalates. When used
as indicated, very limited exposure to trace amounts of phthalates may occur.
There is no clear clinical evidence that this degree of exposure increases
clinical risk. However, in order to minimize risk of phthalate exposure in
children and nursing or pregnant women, this product should only be used as
directed.
WARNING:
The ventilator offers a variety of breath delivery options. Throughout the
patient's treatment, the clinician should carefully select the ventilation mode
and settings to use for that patient based on clinical judgment, the condition
and needs of the patient, and the benefits, limitations and characteristics of
the breath delivery options. As the patient's condition changes over time,
periodically assess the chosen modes and settings to determine whether or
not those are best for the patient's current needs.

Operators Manual

Part numbers

WARNING:
Inspect the Proximal Flow Sensor prior to use, and do not use it if the sensor
body, tubing, or connector are damaged, occluded, or broken.
WARNING:
Do not use the Proximal Flow Sensor if there are kinks in the tubing.
WARNING:
Prior to patient ventilation with the Proximal Flow Option, run SST with the
exact configuration that will be used on the patient. This includes a neonatal
patient circuit, Proximal Flow Sensor, and all accessories used with the patient
circuit. If SST fails any Proximal Flow Sensor test, check the patient circuit and
the Proximal Flow sensor for leaks or occlusions and replace the flow sensor,
if necessary. If SST continues to fail, it may indicate a malfunction or a leak
within the Proximal Flow hardware which could compromise accuracy or
increase the likelihood of cross-contamination; thus, replace the Proximal
Flow hardware.
WARNING:
Changing ventilator accessories can change the system resistance and
compliance. Do not add or remove accessories after running SST.
WARNING:
If the Proximal Flow Option fails to respond as described in this appendix,
discontinue use until correct operation is verified by qualified personnel.
WARNING:
The Proximal Flow Sensor measures gas flow at the patient wye. The actual
volume of gas delivered to the patient may be affected by system leaks
between the patient and the Proximal Flow Sensor, such as a leak that could
occur from the use of an uncuffed endotracheal tube.
WARNING:
Position the Proximal Flow Sensor exactly as described in this appendix or the
Instructions for Use (IFU) provided with the sensor.

Operators Manual

F-3

Proximal Flow Option Appendix

F-4

WARNING:
Do not position the Proximal Flow Sensor cables or tubing in any manner that
may cause entanglement, strangulation or extubation which could lead to
hypercarbia or hypoxemia. Use the cable management clips supplied to
mitigate this risk.
WARNING:
To reduce the risk of extubation or disconnection, do not apply tension to or
rotate the Proximal Flow Sensor by pulling on the Proximal Flow Sensors
tubing.
WARNING:
Do not install the Proximal Flow Sensor in the patient circuit if the sensor is
not also connected to the BDU.
WARNING:
Excessive moisture in the Proximal Flow Sensor tubing may affect the
accuracy of the measurements. Periodically check the sensor and tubing for
excessive moisture or secretion build-up.
WARNING:
The Proximal Flow Sensor is intended for single use only. Do not re-use the
sensor. Attempts to clean or sterilize the sensor may result in
bioincompatibility, infection, or product failure risks to the patient.
WARNING:
Install the Proximal Flow Sensor as shown. Reference Attaching Proximal
Flow Sensor, p. F-14. Improper orientation of the flow sensor could lead to
misinterpretation of data or incorrect ventilator settings.
Caution:
Do not use aerosolized medications with the Proximal Flow Sensor. Such
medications may damage the sensor.
Caution:
To prevent damage to pneumatic lines, use supplied cable management clips.

Operators Manual

Part numbers

F.6

Caution:
Use only Covidien-branded Proximal Flow Sensors with the Proximal Flow
Option.

Proximal Flow Option Description


The Proximal Flow Option measures pressure, flow, and volume at the patient
wye. A Printed Circuit Board Assembly (PCBA) containing the electronics and
pneumatics for the Proximal Flow Option is installed in the ventilator on the
Option Host Card. Data measured by the Proximal Flow Sensor are displayed
on the GUI for monitoring purposes, not for ventilator control. When the ventilator has a Proximal Flow Sensor installed, both proximal flow and proximal
pressure measurements are obtained and displayed on the GUI.
A manual purge control is also provided to clear pneumatic lines for accurate
pressure measurements. When a manual purge is requested, the ventilator will
not allow another purge for at least 30 seconds. Reference Sensor Calibration
and Sensor Line Purging, p. F-8 for more information on the purge function.

F.6.1

Proximal Flow Option components


The Proximal Flow Option consists of the following components:
Proximal Flow Option PCBA Installed on the Option Host Card in the BDU, this
printed circuit board assembly contains a pressure sensor to measure the pressure
difference between the flow sensor lines and the interfaces required to convert
analog measurements from the Proximal Flow Sensor into digital data displayed
by the ventilator. The PCBA also contains valves and an accumulator for purging
the sensor lines from blockages.
Proximal Flow Sensor The Puritan Bennett Proximal Flow Sensor is required for
use with the Proximal Flow Option. The sensor is installed near the patient circuit
wye. The other end of the sensor connects to the ventilators front panel behind
a clear door designed to protect the connection point from exposure to spills or
from sprayed liquids during cleaning and disinfection.

Operators Manual

F-5

Proximal Flow Option Appendix

FigureF-1.Proximal Flow Sensor

F.7

On-screen symbols
When using the Proximal Flow Option, flow, pressure, and volume waveform
data, along with delivered and exhaled volumes are derived from Proximal
Flow Sensor measurements at the patient circuit wye. Proximal flow data are
displayed on the waveform plot with a Y appearing in inverse video next to the
measurement symbol.

F-6

Operators Manual

Part numbers

FigureF-2.Sample GUI screen Showing Proximal Flow Data

Data measured using Proximal Flow Sensor


VY Flow throughout the breath cycle (at patient circuit wye)
PCIRCY Pressure throughout the breath cycle (at patient circuit wye)

Inspired and exhaled flows and volumes at the patient wye are measured and
identified by the symbols shown below, and correspond to their non-proximal
flow equivalents. These values appear in the patient data panel if so configured. Reference Vital Patient Data on page 3-42, and the figure above.

Operators Manual

F-7

Proximal Flow Option Appendix

TableF-2.Proximal Flow Option Patient Data Symbols


Data Symbol
VTIY

Inspired tidal volume (mandatory or spontaneous at patient circuit wye)

VTEY

Exhaled tidal volume (at patient circuit wye)

VTE SPONTY

Exhaled spontaneous tidal volume (at patient circuit wye)

VTE MANDY

Exhaled mandatory tidal volume (at patient circuit wye)

VE TOTY
VY
VTLY

Description

Exhaled total minute volume (at patient circuit wye)


Flow throughout the breath cycle (at patient circuit wye)
Inspired tidal volume (at patient circuit wye with Leak Sync enabled)

Note:
In the patient data symbols shown above, the Y appears in inverse video, as shown.
Reference Sample GUI screen Showing Proximal Flow Data, p. F-7.
Note:
When the Proximal Flow and Leak Sync options are enabled, the following parameters
are available for display:

VTLY and VTL

LEAK and LEAKY

When only the Proximal Flow option is enabled, VTIY and VTI are available for display
When a Y appears in the symbol, the data are measured with the proximal flow
sensor. When a Y is absent from the symbol, the data are measured by the
ventilators internal flow sensors.
F.8

Sensor Calibration and Sensor Line Purging


To ensure accurate pressure and flow measurements, the ventilator performs
an autozero function to calibrate the Proximal Flow Sensor. It does this by periodically opening the pressure sensor on the Proximal Flow Option PCBA to
atmosphere during exhalation, and uses the resulting measurements as offset
corrections.

F-8

Operators Manual

Part numbers

The purge function is designed to clear the pneumatic lines of fluids that may
collect, and is performed periodically by sending a brief flow of air through the
sensor lines. Autozero and purge functions are only active during exhalation
which limits the effect of the purge gas on delivered oxygen concentration.
During the autozero or automatic purge processes, the measurement and
display of proximal flow data is not shown in real time and a brief message
appears on the GUI indicating the purge process is occurring.
During autozero or automatic purge processes, the pressure waveforms, when
shown display the current PEEP value and the flow waveform, when shown,
displays a value of 0.
FigureF-3.Message During Autozero and Purge Processes

Screenshot of Prox purge in progress

F.9

SST Requirements
SST must be run prior to ventilation and all circuit components and accessories
must be installed in the configuration to be used on the patient in order for the
ventilator to calculate the correct compliance and resistance. This includes a
neonatal patient circuit, Proximal Flow Sensor, and other accessories used
during ventilation. Reference To run SST in Chapter 3 of this manual. There is
also a table listing the general SST test sequence located in that section. Ref-

Operators Manual

F-9

Proximal Flow Option Appendix

erence the table below for a listing of the test sequence when running SST
with the Proximal Flow Option.

Note:
Failure of the Proximal Flow Option to pass SST does not prevent ventilation, but will
prevent measurement with the Proximal Flow Option. The ventilator will use its
internal flow sensors for measurement instead of the Proximal Flow Option.
TableF-3.Proximal Flow Option SST test Sequence
Test Step

F-10

Function

Comments

SST Flow Sensor Cross Check

Tests O2 and Air Flow Sensors

N/A

SST EV Performance

Calibrates the exhalation valve


and creates a table for use
during calculations.

N/A

SST Circuit Pressure

Exercises delivery PSOL.


Checks inspiratory and expiratory autozero solenoids.
Cross-checks inspiratory and
expiratory pressure transducers
at various pressures.

N/A

SST Leak

Tests ventilator breathing


system for leaks.

N/A

SST Exhalation Filter

Checks for exhalation filter


occlusion and exhalation compartment occlusion.

Ventilator prompts the user to


block the proximal flow sensor
outlet during Leak test. When
prompted to reconnect the
patient to the expiratory filter
during the Exhalation Filter test,
resume blocking the proximal
flow sensor outlet.

SST Circuit Resistance

Checks for inspiratory and expiratory limb occlusions, and calculates and stores the
inspiratory and expiratory limb
resistance parameters.

N/A

SST Circuit Compliance

Calculates the attached patient


circuit compliance.

N/A

SST Prox

Verifies functionality of Proximal Flow System.

Includes tests of barometric


pressure, autozero, purge, and
pressure cross check functions

Operators Manual

Part numbers

F.9.1

Attaching the Proximal Flow Sensor for SST


During SST the ventilator prompts to attach the Proximal Flow Sensor.
To attach the Proximal Flow Sensor to the patient circuit
1.
Verify the Proximal Flow Sensor, pneumatic lines, and connector are not damaged.
Open the connector panel door and firmly attach the sensor connector to the
receptacle in the BDUs front connector port labeled Prox.

2.

FigureF-4.Attaching Proximal Flow Sensor to Ventilator

Proximal Flow Sensor connector


insertion port

Proximal Flow Sensor connector

3.

When prompted, block the breathing circuit wye.

4.

When prompted to attach the Proximal Flow Sensor, unblock the circuit wye and
insert the smaller end of the sensor into the wye.

5.

When prompted, cap or seal the larger end of the sensor (marked with UP and
an arrow).

6.

Follow the prompts to complete SST.

Operators Manual

F-11

Proximal Flow Option Appendix

If SST fails, check the patient circuit and flow sensor connections for leaks or
occlusions and replace the Proximal Flow Sensor, if necessary. Replace the
Proximal Flow Option hardware if SST continues to fail, then repeat SST to
determine circuit compliance and resistance. Reference the Puritan Bennett
980 Series Ventilator Hardware Options Installation Instructions, p/n
10084704 for instructions on replacing the Proximal Flow Option hardware.
F.10

Disabling/Enabling the Proximal Flow Option


The Proximal Flow Sensor can function in the ENABLED state only if the circuit
type is NEONATAL. Assuming the Proximal Flow Option is available and the
vent type is INVASIVE, the New Patient default value is ENABLED. If the vent
type is NON-INVASIVE, the Proximal Flow Option is DISABLED. The system
becomes ENABLED again if the vent type is changed to INVASIVE.
After SST has been performed, the clinician may disable the Proximal Flow
Option, if desired.
To disable or enable the Proximal Flow Option
1.
In the constant access icons area, touch the configure icon. A menu containing
tabs appears.

F-12

2.

Touch the Options tab. A screen appears containing the Installed Options and Prox
tabs.

3.

Touch Enabled or Disabled to enable or disable the Prox Flow option.

Operators Manual

Part numbers

FigureF-5.Enabling/disabling Proximal Flow Sensor

F.11

Note:
If the Proximal Flow Option has been disabled or enabled, SST does not have to be rerun unless the breathing circuit or other breathing system accessories have been
changed, removed, or added.

Using the Proximal Flow Sensor


Review and follow all warnings prior to patient ventilation with the Proximal
Flow Sensor. Reference Safety Information, p. F-2, and ensure the Proximal
Flow Sensor option is enabled.
To connect the Proximal Flow Sensor to the ventilator
1.
Verify the Proximal Flow Sensor, pneumatic lines, and connector are not damaged
in any way.
2.

Operators Manual

Open the connector panel door and firmly attach the sensor connector to the
right-most receptacle in the BDUs front connector port labeled Prox. Reference
Attaching Proximal Flow Sensor to Ventilator, p. F-11.

F-13

Proximal Flow Option Appendix

To attach the Proximal flow sensor between the endotracheal tube and
patient circuit
1.
Connect the larger end of the sensor (marked with UP and an arrow) to the
endotracheal tube. Reference the figure below. Do not force the connection;
when the sensor is oriented correctly, insertion requires little effort.

Note:
If using a Heat-Moisture Exchanger (HME) on the endotracheal tube, place the
Proximal Flow Sensor between the HME and the breathing circuit wye.
FigureF-6.Attaching Proximal Flow Sensor

F-14

Endotracheal tube

Breathing circuit wye

2.

Connect the smaller end of the sensor to the breathing circuit wye.

3.

Ensure the sensor tubing is positioned in an upward direction, as shown in the


figure above. If the sensor needs repositioning, DO NOT rotate it by pulling on
the tubing. Reposition as follows:
a.

Grasp the sensors plastic body with one hand and the breathing circuit wye
with the other hand.

b.

Rotate the sensor body and wye towards each other until the sensor tubing is
upright.

c.

Confirm a tight connection between the sensor and breathing circuit wye.

Operators Manual

Part numbers

4.

F.11.1

Use the three cable management clips provided with the sensor to attach the
sensor tubing to the breathing circuit tubing. Space the clips evenly along the
length of the sensor tubing. Twist the ends of each clip to close.

Note:
When the ventilator is set up for Proximal Flow Option operation, the Proximal
Flow Sensor can be switched as necessary. There is no need to run SST after
switching sensors unless the breathing circuit or other ventilator accessories have
been changed.

How to Perform a Manual Purge


A manual purge may be performed any time the sensor lines contain excessive
condensation, moisture, or secretions.
To perform a manual purge
1.
Touch the Configure icon on the in the constant access icons area of the GUI.
2.

Touch the Options tab. A screen appears containing the Installed Options and Prox
tabs.

3.

Touch the Prox tab. The Prox Setup screen appears.

4.

Touch Start that appears next to the text Prox Manual Purge: To begin touch the
Start button. During the purge, a message appears in the GUI prompt area
stating the purge process is being performed.Reference Message During Autozero
and Purge Processes, p. F-9.

Operators Manual

F-15

Proximal Flow Option Appendix

FigureF-7.Manual Purge

F.12

Alarms
If the Proximal Flow Option becomes inoperable during ventilation, the ventilator annunciates an alarm and flow sensing reverts to the ventilators internal
delivery and exhalation valve flow sensors. This switch over may be triggered
by any of the following events:

The Proximal Flow Sensor is not detected

Pressure and flow readings are out of range

Hardware problems are reported by the Proximal Flow Option PCBA

There is a communication failure between the ventilator and the Proximal Flow
option

If any of these conditions occur, the GUI displays an alarm message similar the
one shown below.Follow the information contained in the remedy message to
troubleshoot the alarm.

F-16

Operators Manual

Part numbers

FigureF-8.Alarm Message Prox Inoperative

F.13

Ranges, Resolutions, and Accuracies


Reference Patient Data Range and Resolution in Chapter 11 of this manual for
Proximal exhaled tidal volume, Proximal inspired tidal volume, Proximal
exhaled minute volume, and Proximal flow patient data parameters.

F.13.1

Proximal Flow Sensor Specifications


TableF-4.Proximal Flow Sensor Volume Accuracy
Measurement

Accuracy1

Exhaled tidal volume

(1.0 mL + 10% of reading)

Inspired tidal volume

(1.0 mL + 10% of reading)

1. The conditions under which the accuracy values apply are as follows:
Sensor is used as described in this appendix and/or the Instructions for Use provided with the sensor

Operators Manual

F-17

Proximal Flow Option Appendix

TableF-5.Proximal Flow Sensor Specifications

F.14

Parameter

Specification

Weight

6.6 g

Dead space

< 1 mL

Pressure drop

1.5 cmH2O at 10 L/min

Part Numbers
The table below lists the part numbers for the Proximal Flow Option Kit and
individual components.
TableF-6.Proximal Flow Option Component Part Numbers
Item

Proximal Flow Option Kit

Part Number
10084331

Includes:
Installation hardware and accessories

F-18

Proximal Flow Sensor, Neonatal (package of 10)


NOTE: Includes 3 cable management clips

10047078

Proximal Flow Sensor module

10087622

Interconnect PCBA

10083941

Purge Control Cable

10083940

Purge Supply Line

10083966

PCBA Mounting Screws

10083963

Proximal Flow Option Label

10005748

Operators Manual

Glossary

TableGlossary-1.Glossary of Ventilation Terms


analysis message

A message displayed on the GUI screen during an alarm condition, identifying the root cause of the alarm.

assist breath

A mandatory breath triggered by patient inspiratory effort in A/C and


SIMV modes.

assist-control A/C mode

A ventilation mode where only mandatory VC, PC, or VC+ breaths are
delivered to the patient.

augmented alarm

The initial cause of an alarm has precipitated one or more related alarms.
When an alarm occurs, any subsequent alarm related to the cause of this
initial alarm augments the initial alarm.

autotriggering

The ventilator delivers repeated, unintended breaths triggered by fluctuating flows or pressures as opposed to patient demand. Patient circuit
leaks and low flow or pressure sensitivity settings are common causes of
autotriggering.

background checks

Continuously running tests during ventilation that assess the ventilators


electronics and pneumatics hardware.

backup ventilation
(BUV)

A safety net feature which is invoked if a system fault in the mix subsystem, inspiratory subsystem, or expiratory subsystem occurs compromising the ventilators ability to ventilate the patient as set.

base flow

A constant flow of gas through the patient circuit during the latter part
of exhalation during flow triggering (V-TRIG). The value of this base flow
is 1.5 L/min greater than the operator selected value for flow sensitivity.

base message

A message given by the ventilator during an alarm condition, identifying


the alarm.

batch changes

Changes to multiple settings that go into effect at the same time.

battery back-up system

The system for supplying battery back-up power to a device. The ventilator's battery back-up system consists of a single primary battery to
provide up to one (1) hour of battery power to the ventilator. An optional
extended battery with the same characteristics as the primary battery is
available.

BD, BDU

Breath delivery or breath delivery unit. The ventilator component that


includes inspiratory and expiratory pneumatics and electronics.

Glossary-1

TableGlossary-1.Glossary of Ventilation Terms (Continued)

Glossary-2

BiLevel mode

A mixed ventilation mode combining mandatory and spontaneous


breaths, where two levels of pressure are delivered (PL and PH) corresponding to expiratory and inspiratory times TL and TH.

BOC

British Oxygen Company. A standard for high pressure gas inlet fittings.

breath stacking

The delivery of a second inspiration before the previous exhalation is


complete.

BTPS

Body temperature and pressure, saturated, 37C, at ambient barometric


pressure, at 100% relative humidity.

cmH2O

Centimeters of water. A unit of pressure approximately equal to 1 hPa.

compliance volume

The volume of gas that remains in the patient circuit and does not enter
the patient's respiratory system.

constant during rate


change

One of three breath timing variables (inspiratory time, I:E ratio, or expiratory time) the operator can hold constant when the respiratory rate
setting changes. Applies only to the pressure control (PC) mandatory
breath type (including VC+ and BiLevel).

control breath

A ventilator-initiated mandatory breath delivered in A/C mode

CPU

Central processing unit. The electronic components of the ventilator (BD


and GUI) responsible for interpreting and executing instructions entered
by the operator.

dependent alarm

An alarm that arises as a result of another primary alarm (also referred to


as an augmentation).

DSENS

Disconnect sensitivity. A setting that specifies the allowable loss (percentage) of delivered tidal volume, which if equaled or exceeded, causes the
ventilator to declare a DISCONNECT alarm. The greater the setting, the
more returned volume must be lost before DISCONNECT is detected. If
the Leak Sync option is in use, DSENS is the maximum allowable leak rate
and is expressed in terms of L/min.

DISS

Diameter index safety standard. A standard for high pressure gas inlet fittings.

ESENS

Expiratory sensitivity. A setting that determines the percent of peak inspiratory flow (or flow rate expressed in L/min in a PAV breath) at which the
ventilator cycles from inspiration to exhalation for spontaneous breaths.
Low settings will result in longer spontaneous inspirations. In an IE Sync
triggered breath, when the PSYNC signal drops below the ESENS setting,
the ventilator cycles to exhalation.

EST

Extended self test. A comprehensive test of ventilator function, intended


to be run by qualified service personnel.

Operators Manual

TableGlossary-1.Glossary of Ventilation Terms (Continued)


expiratory pause

an operator-initiated maneuver that closes the inspiration (proportional


solenoid) and exhalation valves during the exhalation phase of a mandatory breath. The maneuver can be used to determine intrinsic (auto) PEEP
(PEEPI).

exhalation valve (EV)

The valve in the expiratory limb of the ventilator breathing system that
controls PEEP.

f, fTOT

Respiratory rate, as a setting (f) in A/C, SIMV, and BiLevel the minimum
number of mandatory breaths the patient receives per minute. As a monitored value (fTOT), the average total number of breaths delivered to the
patient.

FAILURE

A category of condition detected during SST or EST that causes the ventilator to enter the safety valve open state. A ventilator experiencing a
FAILURE requires removal from clinical use and immediate service.

flow pattern

A setting that determines the gas flow pattern of mandatory volumecontrolled breaths.

gold standard test


circuit

Test circuit designed for use with EST.

GUI

Graphical user interface. The ventilators touch screen used to enter


patient settings. and alarm settings, including off-screen keys, soft keys,
and knobs.

hard bound

A ventilator setting that has reached its minimum or maximum limit.

high-priority alarm

As defined by international standards organizations, an alarm that


requires immediate attention to ensure patient safety. When a high-priority alarm is active, the red high-priority LED indicator flashes and the
high-priority audible alarm sounds (a repeating sequence of five tones
that repeats twice, pauses, then repeats again), and the alarm banner on
the GUI screen shows an alarm message with the ( !!! ) symbol.

HME

Heat-moisture exchanger. A humidification device, also called an artificial


nose.

hPa

Hectopascal. A unit of pressure, approximately equal to 1 cmH2O.

humidification type

A setting for the type of humidification system (HME, non-heated expiratory tube, or heated expiratory tubing) in use on the ventilator.

I:E ratio

The ratio of inspiratory time to expiratory time. Also, the operator- set
timing variable that applies to PC and VC+ mandatory breaths.

inspiratory pause

An operator-initiated maneuver that closes the inspiration (proportional


solenoid) and exhalation valves at the end of the inspiratory phase of a
mandatory breath. The maneuver can be used to determine static compliance (CSTAT) and static resistance (RSTAT).

Operators Manual

Glossary-3

TableGlossary-1.Glossary of Ventilation Terms (Continued)

Glossary-4

invasive ventilation

Patient ventilation while intubated with an endotracheal (or tracheostomy) tube.

kPa

Kilopascal. A unit of pressure approximately equal to 10 cmH2O.

latched alarm

An alarm whose visual alarm indicator remains illuminated after the


alarm has autoreset.

L/min

Liters per minute. A unit of flow.

low-priority alarm

As defined by international standards organizations, an alarm that indicates a change in the patient-ventilator system. During a low-priority
alarm, the yellow low-priority LED indicator lights, the low-priority
audible alarm (one tone) sounds, and the GUI screen shows an alarm
banner with the ( ! ) symbol.

lockable alarm

An alarm that does not terminate an active alarm silence function.

maintenance

All actions necessary to keep equipment in, or restore it to, serviceable


condition. Includes cleaning, servicing, repair, modification, overhaul,
inspection, and performance verification.

mandatory breath

A breath whose settings and timing are preset; can be triggered by the
ventilator, patient, or operator.

mandatory type

The type of mandatory breath: volume control (VC), VC+, or pressure


control (PC).

manual inspiration

An operator-initiated mandatory (OIM) breath.

medium-priority alarm

As defined by international standards organizations, an abnormal condition that requires prompt attention to ensure the safety of the patient.
When a medium-priority alarm is active, the yellow medium-priority LED
indicator flashes, the medium- priority audible alarm (a repeating
sequence of three tones) sounds, and the GUI screen shows an alarm
banner with the ( !! ) symbol.

mode

Ventilatory mode. The algorithm that determines type and sequence of


breath delivery.

NIST

Non-interchangeable screw thread. A standard for high pressure gas inlet


fittings.

non-invasive ventilation (NIV)

Patient ventilation without the use of an endotracheal tube; instead


using interfaces such as masks, nasal prongs, or uncuffed endotracheal
tubes.

non-technical alarm

An alarm caused due to a fault in the patient-ventilator interaction or a


fault in the electrical or gas supplies that the practitioner may be able to
alleviate.

normal ventilation

The state of the ventilator when breathing is in progress and no alarms


are active.

Operators Manual

TableGlossary-1.Glossary of Ventilation Terms (Continued)


O2%

Both a ventilator setting and a monitored variable. The O2% setting


determines the percentage of oxygen in the delivered gas. The O2%
monitored data is the percentage of oxygen in the gas delivered to the
patient, measured at the ventilator outlet upstream of the inspiratory
filter.

OIM

Operator-initiated mandatory breath. A breath delivered when the operator presses the MANUAL INSP key.

ongoing background
checks

Continuously running tests during ventilation that assess the ventilator's


electronics and pneumatics hardware.

OSC

Occlusion status cycling. A state invoked during a severe occlusion. In this


mode, the ventilator periodically attempts to deliver a pressure-based
breath while monitoring the inspiration and expiration phases for the
continuing existence of the occlusion.

OVERRIDDEN

The final status of an SST or EST run in which the operator used the override feature. (The ventilator must have ended the test with an ALERT condition.) Failures cannot be overridden.

patient circuit

The entire inspiratory-expiratory conduit, including tubing, humidifier,


and water traps.

patient data alarm

An alarm condition associated with an abnormal condition of the


patient's respiratory status.

patient problems

A definition used by the ventilator's safety net. Patient problems are


declared when patient data are measured equal to or outside of alarm
thresholds and are usually self-correcting or can be corrected by a practitioner. The alarm monitoring system detects and announces patient
problems. Patient problems do not compromise the ventilator's performance.

PBW

Predicted body weight, a ventilator setting that specifies the patient's


body weight assuming normal fat and fluid levels. Determines absolute
limits on tidal volume and peak flow, and allows appropriate matching
of ventilator settings to the patient.

PC

Pressure control. A mandatory breath type in which the ventilator delivers


an operator-set inspiratory pressure for an operator- set inspiratory time.
Available in A/C and SIMV modes, and for operator-initiated mandatory
(OIM) breaths in SPONT mode.

PE

Expiratory pressure transducer.

PEEP

Positive end expiratory pressure. The measured circuit pressure (referenced to the patient wye) at the end of the expiratory phase of a breath.
If expiratory pause is active, the displayed value reflects the level of any
active lung PEEP.

Operators Manual

Glossary-5

TableGlossary-1.Glossary of Ventilation Terms (Continued)


PEEPI

Intrinsic PEEP. Indicates a calculated estimate of the pressure above the


PEEP level at the end of exhalation. Determined during an expiratory
pause maneuver.

PI

Inspiratory pressure. The operator-set inspiratory pressure at the patient


wye (above PEEP) during a pressure control (PC) mandatory breath.

PI

Inspiratory pressure transducer.

PI END

End inspiratory pressure. The pressure at the end of the inspiration phase
of the current breath. If plateau is active, the displayed value reflects the
level of end-plateau pressure.

PIM

Patient-initiated mandatory breath. A mandatory breath triggered by


patient inspiratory effort.

PMEAN

Mean circuit pressure, a calculation of the measured average patient


circuit pressure over an entire respiratory cycle.

PPEAK

Maximum circuit pressure, the maximum pressure during the inspiratory


and expiratory phases of a breath.

primary alarm

An initial alarm.

PS

Pressure support, a spontaneous breath type in which the ventilator


delivers an operator-set pressure (in addition to PEEP) during the inspiratory phase. Available in SPONT, SIMV, and BiLevel modes.

PSENS

Pressure sensitivity. The operator-set pressure drop below PEEP (derived


from the patient's inspiratory flow) required to begin a patient-initiated
breath when pressure triggering is selected.

PSOL

Proportional solenoid valve.

PSUPP

Pressure support. A setting of the level of inspiratory assist pressure


(above PEEP) at the patient wye during a spontaneous breath (when
spontaneous breath type is PS).

P-TRIG

Pressure triggering, a method of recognizing patient inspiratory effort in


which the ventilator monitors pressure in the patient circuit. The ventilator triggers a breath when the airway pressure drops by at least the value
selected for pressure sensitivity
(PSENS).

Glossary-6

remedy message

A message displayed on the GUI during an alarm condition suggesting


ways to resolve the alarm.

resistance

The flow-dependent pressure drop across a conduit. Measured in


cmH2O/L/s or hPa/L/s.

Operators Manual

TableGlossary-1.Glossary of Ventilation Terms (Continued)


restricted phase of
exhalation

The time period during the exhalation phase where an inspiration trigger
is not allowed. The restricted phase of exhalation is defined as the first
200 ms of exhalation, OR the time it takes for expiratory flow to drop to
50% of the peak expiratory flow, OR the time it takes for the expiratory
flow to drop to 0.5 L/min (whichever is longest). The restricted phase
of exhalation will end after five (5) seconds of exhalation have elapsed
regardless of the measured expiratory flow rate.

rise time %

A setting that determines the rise time to achieve the set inspiratory pressure in pressure-controlled (PC), VC+, BiLevel, or pressure-supported (PS)
breaths. The larger the value, the more rapid the rise of pressure.

safety net

The ventilator's strategy for responding to patient problems and system


faults.

safety valve (SV)

A valve residing in the ventilators inspiratory module designed to limit


pressure in the patient circuit. When open, it allows the patient to
breathe room air (if able to do so).

safety ventilation

A mode of ventilation active if the patient circuit is connected before ventilator startup is complete, or when power is restored after a loss of five
(5) minutes or more.

service mode

A ventilator mode providing a set of services tailored to the needs of


testing and maintenance personnel. When in the service mode, the ventilator does not provide ventilation.

SIMV

Synchronous intermittent mandatory ventilation. A ventilatory mode in


which the ventilator delivers one mandatory breath per breath cycle and
as many spontaneous breaths as the patient can trigger during the
remainder of the breath cycle.

SIS

Sleeved index system. A standard for high pressure gas inlet fittings.

soft bound

A ventilator setting that has reached its recommended high or low limit,
accompanied by an audible tone. Setting the ventilator beyond this limit
requires the operator to acknowledge a visual prompt to continue.

SPONT

Spontaneous. A ventilatory mode in which the ventilator delivers only


spontaneous breaths. In SPONT mode, the patient triggers all breaths
delivered by the ventilator with no set mandatory respiratory rate. The
patient controls the breath variables, potentially augmented by support
pressure.

spontaneous type

A setting that determines whether spontaneous breaths are pressuresupported (PS), tube-compensated (TC), volume-supported (VS), or proportionally assisted (PAV).

SST

Short self test. A test that checks circuit integrity, calculates circuit compliance and filter resistance, and checks ventilator function. Operator
should run SST at specified intervals and with any replacement or alteration of the patient circuit.

Operators Manual

Glossary-7

TableGlossary-1.Glossary of Ventilation Terms (Continued)

Glossary-8

STPD

Standard temperature and pressure, dry. Defined as dry gas at a standard


atmosphere (760 mmHg, 101.333 kPa, approximately 1.0 bar) and 0C.

SVO

Safety valve open. An emergency state in which the ventilator opens the
safety valve so the patient can breathe room air unassisted by the ventilator (if able to do so). An SVO state does not necessarily indicate a ventilator inoperative condition. The ventilator enters an SVO state if a
hardware or software failure occurs that could compromise safe ventilation, with the loss of the air and oxygen supplies, or if the system detects
an occlusion.

system fault

A definition used by the ventilator's safety net. System faults include


hardware faults (those that originate inside the ventilator and affect its
performance), soft faults (faults momentarily introduced into the ventilator that interfere with normal operation), inadequate supply (AC power
or external gas pressure), and patient circuit integrity (blocked or disconnected circuit).

TA

Apnea interval, the operator-set variable that defines the breath-tobreath interval which, if exceeded, causes the ventilator to declare apnea
and enter apnea ventilation.

Tb

Breath time cycle during mechanical ventilation.

TE

Expiratory time. The expiratory interval of a breath. Also the operator-set


timing variable that determines the expiratory period for pressure-controlled (PC) or VC+ mandatory breaths.

technical alarm

An alarm occurring due to a violation of any of the ventilator's self monitoring conditions, or detected by background checks.

TI

Inspiratory time, the inspiratory interval of a breath. Also, the operatorset timing variable that determines the inspiratory interval for pressurecontrolled (PC) or VC+ mandatory breaths.

Tm

Mandatory interval portion of SIMV breath cycle; it is reserved for a PIM.

TPL

Plateau time. The amount of time the inspiration phase of a mandatory


breath is extended after inspiratory flow has ceased and exhalation is
blocked. Increases the residence time of gas in the patient's lungs.

Ts

Spontaneous interval portion of SIMV breath cycle; it is reserved for spontaneous breathing throughout the remainder of the breath cycle.

VE TOT

Minute volume, the expiratory tidal volume normalized to unit time


(L/min). The displayed value is compliance- and BTPS-compensated.

VBS

Ventilator breathing system. Includes the gas delivery components of the


ventilator the patient circuit with tubing, filters, humidifier, and other
accessories; and the ventilator's expiratory metering and measurement
components.

Operators Manual

TableGlossary-1.Glossary of Ventilation Terms (Continued)


VC

Volume control, a mandatory breath type in which the ventilator delivers


an operator-set tidal volume, peak flow, and flow pattern. Available in
A/C and SIMV modes, and for operator-initiated mandatory (OIM)
breaths in SPONT mode.

Ventilation Assurance

A feature on the 980 Series Ventilator which enables ventilation to continue when a critical system error occurs, by entering the Backup Ventilation (BUV) state.

Ventilator Inoperative
(vent inop)

An emergency state the ventilator enters if it detects a hardware failure


or a critical software error which could compromise safe ventilation.
During a ventilator inoperative condition, the safety valve opens to allow
the patient to breathe room air (if able to do so) unassisted by the ventilator. Qualified service personnel must power up the ventilator and run
EST before normal ventilation can resume.

VIM

Ventilator-initiated mandatory breath. A breath that is delivered at a time


determined by the ventilator

VMAX

Peak flow. A setting of the peak (maximum) flow of gas delivered during
a VC mandatory breath. (Combined with tidal volume, flow pattern, and
plateau, constant peak flow defines the inspiratory time.) To correct for
compliance volume, the ventilator automatically increases the peak flow.

VSENS

Flow sensitivity. A setting that determines the rate of flow inspired by the
patient that triggers the ventilator to deliver a mandatory or spontaneous
breath (when flow triggering is selected).

VT

Tidal volume. A setting that determines the volume inspired and expired
with each breath. The VT delivered by some Puritan Bennett ventilators is
an operator-set variable that determines the volume delivered to the
patient during a mandatory, volume-based breath. VT is compliancecompensated and corrected to body temperature and pressure, saturated (BTPS).

V-TRIG

Flow triggering. A method of recognizing patient inspiratory effort in


which the ventilator monitors the difference between inspiratory and
expiratory flow measurements. The ventilator triggers a breath when the
difference between inspiratory and expiratory flows increases to a value
that is at least the value selected for flow sensitivity (VSENS.

TableGlossary-2.Units of Measure

Operators Manual

cm

Centimeter. A unit of length.

ft

Feet. A unit of length.

Hz

Hertz. A unit of frequency, indicating cycles per second.

kg

Kilogram. A unit of weight

Glossary-9

TableGlossary-2.Units of Measure (Continued)


L

Liter. A unit of volume

lb

Pound. A unit of weight.

Meter. A unit of length.

mL

Milliliter. A unit of volume.

ms

Millisecond. A unit of time.

Second. A unit of time

Volts. A unit of voltage

VA

Volt-amperes. A unit of power.

TableGlossary-3.Technical Abbreviations

Glossary-10

AC, also ac

Alternating current. The movement of electrical charge that periodically


reverses direction.

ASCII

American Standard Code for Information Interchange. A standard character encoding scheme.

CE

A certification mark issued under the authority of the European


Common Market that indicates compliance with the Medical Device
Directive, 93/42/EEC.

CSA

Canadian Standards Association.

CRC

Cyclic Redundancy Check or Code. An algorithm or a computational


result based on the remainder of a division defined over the ring of
polynomials in the Galois field GF(2). CRC algorithms are the basis for
data integrity checks.

DC, also dc

Direct current. The movement of electrical charge flowing in a single


direction.

EMC

Electromagnetic compatibility.

EN

European norm (referring to the European Common Market).

ETO

Ethylene oxide.

IEC

International Electrotechnical Commission. A standards organization.

ISO

International Standards Organization. A standards organization.

LCD

Liquid crystal display. A type of visual equipment-operator Interface.

LED

Light-emitting diode. A means of providing visual indications.

MRI

Magnetic resonance imaging.

Operators Manual

TableGlossary-3.Technical Abbreviations (Continued)

Operators Manual

NVRAM, also NovRam

Non-volatile random access memory. Memory that is kept active across


resets and power cycles and is not normally initialized at startup.

POST

Power-on self-test. Software algorithms to verify the integrity of application software and the hardware environment. Power-on self-test
generally occurs at power on, after power loss, or when the device
resets due to a detected fault.

RAM

Random access memory.

Glossary-11

Page Left Intentionally Blank

Glossary-12

Operators Manual

A
A/C mode . . . . . . . . . . . . . 10-3110-34
AC power operation . . . . . . . . . . . . 3-2
accessory
compatibility 9-1
part numbers 9-4
adjusting waveform layout . . . . . 3-45
alarm
AC POWER LOSS 6-35
Apnea 6-35
CIRCUIT DISCONNECT 6-35
dependent 6-4
DEVICE ALERT 6-36
High circuit pressure 6-36
High delivered O2% 6-37
High exhaled minute volume 6-38
High exhaled tidal volume 6-38
High inspired tidal volume 6-38
High respiratory rate 6-39
how to test 6-9
INSPIRATION TOO LONG 6-39
latched 6-5
lockable 6-6
Loss of power 6-36
Low circuit pressure 6-39
Low delivered O2% 6-40
Low exhaled mandatory tidal volume 6-41
Low exhaled spontaneous tidal volume 641
Low exhaled total minute volume 6-41
non-technical 6-18
primary 6-4
prioritization 6-16
PROCEDURE ERROR 6-42
reset 6-8
silence 6-8
symbols 6-6
technical 6-17
volume 6-8
Alarm Functions . . . . . . . . . . . . . . . 6-4
alarm settings range, resolution, accuracy
11-19
alarm settings range, resolution, accuracy
11-17
Apnea Ventilation . . . . . . 10-4110-45
apnea ventilation . . . . . . . . . . . . 10-57
Apnea Ventilation in SIMV . . . . . 10-44
B
Background Diagnostic System . 10-74
battery
life 3-25
battery installation . . . . . . . 3-183-22
BDU indicators
audible 2-35
visual 2-27
BiLevel 2.0 option . . . . . . . . . . . . . .A-1

Breath Delivery Unit . . . . . . . . . . . 2-25


Breath triggers
flow 10-6
IE Sync 10-8
operator-initiated 10-8
pressure 10-5
time-cycled 10-8
BUV settings . . . . . . . . . . . . . . . . 10-75
C
CircuitType and PBW . . . . . . . . . 10-58
compliance compensation 10-1310-17
Compliance Compensation in Volume-based
Breaths . . . . . . . . . . . . . 10-13
Component Cleaning and Disinfection 7-6
Component Sterilization . . . . . . . 7-23
configurable features
alarm volume 3-41
date and time 3-38
large font patient data 3-43
mL/kg ratio 3-40
new patient setup defaults 3-40
O2% 3-40
patient data 3-42
PBW 3-40
pressure units 3-39
screen brightness and keyboard backlight
3-39
screen opacity 3-45
waveforms 3-44
Connect the Ventilator to AC Power 3-5
Connecting the Gas Supplies . . . . . 3-7
Connecting the Patient Circuit . . 3-14
Connectivity to External Patient Monitoring
Systems . . . . . . . . . . . . . . 5-21
Constant Timing Variable for Rate Changes
4-20
Covidien Technical Services
list of International Service Centers 1-15
phone number 1-15
Solv-IT Center Knowledge Base 1-18
Covidien URL . . . . . . . . . . . . . .1-1, 1-15
D
Detecting Occlusion and Disconnect 10-45
10-49
Disconnect . . . . . . . . . . . . . . . . . . 10-47
Disconnect Sensitivity (DSENS) . . 10-70
Display
brightness adjustment 4-5
lock 4-6
E
EMC
compatibility 1-20
emissions 11-30
immunity 11-31

recommended separation distances 11-34


exhalation
airway pressure method 10-9
high circuit pressure limit (backup
method) 10-12
high inspired tidal volume limit (backup
method) 10-12
high ventilator pressure limit (backup
method) 10-12
percent peak flow method 10-10
time cycling method 10-11
time limit (backup method) 10-12
Exhalation Detection and Initiation 10-9
10-12
expiratory module
EVQ removal, disinfection, reassembly 7127-23
Expiratory Pause . . . . . . . 10-5310-54
Expiratory Pause Maneuvers . . . . 4-31
Expiratory Sensitivity (ESENS) . . . . 10-69
Expiratory Time (TE) . . . . . . . . . . 10-66
extended battery installation . . . . 3-22
Extended Self Test (EST) . . . . . . . 10-77
F
Filter Installation . . . . . . . . . . . . . . 3-10
Flow Pattern . . . . . . . . . . . . . . . . 10-63
Flow Sensitivity (VSENS) . . . . . . . . 10-63
G
gas failure cross flow . . . . . . . . . . . 3-8
Gestures
4-64-7
double-tap 4-7
drag and drop 4-7
touch and hold 4-7
gestures
Drag 4-7
swipe 4-6
Graphical User Interface (GUI) . . . 2-16
GUI Control Keys . . . . . . . . . . . . . . 2-16
GUI control keys . . . . . . . . . . . . . . 2-16
GUI indicators
audible 2-23
visual 2-172-20
GUI screen capture . . . . . . . . . . . . . 5-2
H
hard bound . . . . . . . . . . . . . . . 4-3, 4-4
High Pressure (BiLevel) . . . . . . . . 10-66
High Spontaneous Inspiratory Time Limit (2TI
SPONT) . . . . . . . . . . . . . . . 10-70
High Time (BiLevel) . . . . . . . . . . . 10-66
how to enter Service mode . . . . . 3-33
How to Install Accessories . . 3-183-30
how to use the ventilator system 4-74-21

how to use the ventilators user interface 42 . . . . . . . . . . . . . . . . . . . . 4-7


how to view ventilator logs . . . . . . 8-3
Humidification Type . . . . . . . . . . 10-71
humidifier installation . . . . . . . . . 3-27
Humidifier Volume . . . . . . . . . . . 10-71
I
I:E Ratio . . . . . . . . . . . . . . . . . . . . 10-66
icons
configure 2-19
elevate O2 2-19
grid lines 2-20
help 2-19
high priority alarm 2-20
logs 2-18
low priority alarm 2-20
maximize waveform 2-20
medium priority alarm 2-20
pause 2-19
restore waveform 2-20
screen capture 2-19
unread items 2-19
ventilator setup 2-18
waveform layout 2-19
IE Sync option . . . . . . . . . . . . . . . . . C-1
IEC classification . . . . . . . . . . . . . . . 2-3
Inspiration Detection and Initiation 10-4
10-8
Inspiratory Pause . . . . . . . 10-5010-53
Inspiratory pause maneuvers 4-30, 10-50
Inspiratory Pressure (PI) . . . . . . . 10-64
Inspiratory Time (TI) . . . . . . . . . . 10-65
Installation Testing (testing prior to
ventilating a patient) 3-463-51
L
Leak Sync option . . . . . . . . . . . . . . B-1
Low Pressure (BiLevel) . . . . . . . . 10-66
Low Time (BiLevel) . . . . . . . . . . . 10-67
M
Mandatory Breath Delivery . . . . 10-17
Manual Inspiration . . . . . . .4-28, 10-22
Manufacturer . . . . . . . . . . . . . . . . 1-20
MISCA response . . . . . . . . . . . . . . . 5-6
MISCF response . . . . . . . . . . . . . . . 5-10
Mode and Breath Type . . . . . . . . 10-60
Monitored Patient Data . . . . . . . . 6-42
N
NeoMode 2.0 option . . . . . . . . . . . E-1
NIV
alarm settings 4-27
apnea settings 4-27

high spontaneous inspiratory time limit


setting 4-26
setup 4-23
Non-invasive ventilation (NIV) 4-224-28

RS-232 commands
RSET 5-5
SNDA 5-6
SNDF 5-10

O
Occlusion . . . . . . . . . . . . . . . . . . . 10-45
Omni-directional LED . . . . . . . . . . 3-30
On-screen Help . . . . . . . . . . . . . . . 1-18
On-screen Symbols and Abbreviations 2-20
2-23
Operation Verification . . . . . . . . . 3-58
oxygen sensor
calibration 4-34
calibration test 4-34
function 4-32

S
Safety Net . . . . . . . . . . . . . . . . . . 10-71
Serial commands . . . . . . . . . . . . . . . 5-5
serial number interpretation . . . . 1-19
settings
alarm 4-16
apnea 4-15
return to previous 4-20
ventilator 4-10
Short Self Test (SST) . . . . . . . . . . 10-77
SIMV . . . . . . . . . . . . . . . . . . . . . . 10-34
SIMV . . . . . . . . . . . . . . . . . . . . . . 10-39
soft bound . . . . . . . . . . . . . . . . . . . 4-3
Spontaneous (SPONT) Mode 10-3910-41
Spontaneous Breath Delivery . . 10-22
SST
how to run 3-48
outcomes 3-51
results 3-50
test sequence 3-48
Status Display . . . . . . . . . . . . 2-282-34
Storage for Extended Periods . . . 7-26
Surface Cleaning of Ventilator Exterior
Surfaces . . . . . . . . . . . . . . . 7-4
symbols
BDU rear panel label symbols and
descriptions 2-11
safety symbol definitions 1-3
shipping label symbols and descriptions 12
System Related Problems . . . . . . 10-73

P
P0.1 maneuver . . . . . . . . . . . . . . . 10-55
Patient Data Parameters . . . 6-436-50
patient data range and resolution 11-191126
PAV+ option . . . . . . . . . . . . . . . . . .D-1
Peak Inspiratory Flow (VMAX) . . . 10-62
PEEP . . . . . . . . . . . . . . . . . . . . . . . 10-67
PEEP restoration . . . . . . . . . . . . . 10-67
Percent Support (PAV+) . . . . . . . 10-68
Percent Support (TC) . . . . . . . . . . 10-68
Plateau Pressure (PPL) . . . . . . . . . . 6-46
Plateau Time (TPL) . . . . . . . . . . . . 10-63
Pneumatic Diagram . . . . . . . . . . . . 2-36
Power On Self Test (POST) . . . . . 10-76
Preparing the Ventilator for Use . 3-37
Pressure Sensitivity (PSENS) . . . . . 10-64
Pressure Support (PSUPP) . . . . . . . 10-68
primary battery installation . . . . . 3-19
Primary display . . . . . . . . . . . . . . . 2-16
Product Assembly . . . . . . . . . . . . . . 3-2
proximal flow option . . . . . . . . . . . F-1
pushpin . . . . . . . . . . . . . . . . . . . . . . 4-5
Q
quick start use . . . . . . . . . . . . . . . . . 4-8
R
respiratory maneuvers
expiratory pause maneuver 10-53
inspiratory pause maneuver 10-50
NIF maneuver 10-54
respiratory mechanics maneuvers
Negative Inspiratory Force maneuver (NIF)
10-54
P0.1 maneuver 10-55
vital capacity maneuver (VC) 10-56
Respiratory Rate (f) . . . . . . . . . . . 10-61
Rise Time % . . . . . . . . . . . . . . . . . 10-69

T
TC
alarms 10-28
monitored patient data 10-28
PBW and tube ID 10-28
technical description 10-27
tube type, tube ID, humidification 4-14
technical assistance . . . . . . . . . . . 1-15
Technical Services . . . . . . . . . . . . . 1-15
TH:TL ratio (BiLevel) . . . . . . . . . . 10-67
Tidal Volume . . . . . . . . . . . . . . . . 10-62
Tube compensation (TC) . . . . . . 10-27
U
Used Part Disposal . . . . . . . . . . . . . 8-1
Using Battery Power . . . . . . . . . . . . 3-3
V
VC+
maximum pressure adjustments 10-21

startup 10-2010-21
ventilating a new patient . . . . . . . . 4-9
ventilating the same patient . . . . . 4-9
ventilator
alarm log 8-2
available languages 5-1
BDU controls and indicators 2-25
BDU front view 2-8
BDU rear label symbols and descriptions 2112-13
BDU rear view 2-10
BDU right side view 2-14, 2-15
Components List 2-5
connectors 2-35
Description 2-2
EST/SST status log 8-3
function 4-1
gas flow overview 10-2
general event log 8-3
GUI front view 2-6
GUI rear view 2-7
Indications For Use 2-4
Operation 4-74-21
patient data log 8-2
service log 8-3
settings log 8-2
system diagnostic log 8-3
Ventilator Logs . . . . . . . . . . . . . . . . 8-2
Ventilator Operating Modes 3-303-34
ventilator operating modes
Normal mode 3-30
Quick Start mode 3-30
Service mode 3-33
Stand-By state 3-31
Ventilator Protection Strategies 4-354-37
ventilator settings
apnea ventilation 10-57
circuit type and PBW 10-59
configuration 3-373-46
disconnect sensitivity (DSENS) 10-70
DSENS 10-70
ESENS 10-70
flow pattern 10-63

flow sensitivity (VSENS) 10-64


high inspiratory time limit (2TI SPONT) 1071
high pressure (BiLevel) 10-66
high time (BiLevel) 10-66
humidification type 10-71
humidifier volume 10-71
I:E ratio 10-66
inspiratory pressure (PI) 10-65
inspiratory time (TI) 10-65
low pressure (BiLevel) 10-66
low time (BiLevel) 10-67
mode and breath type 10-61
peak inspiratory flow 10-62
peak inspiratory flow (VMAX) 10-62
PEEP 10-67
PEEP restoration 10-67
percent support (PAV+) 10-68
percent support (TC) 10-68
plateau time (TPL) 10-63
pressure sensitivity (PSENS) 10-64
pressure support 10-68
pressure support (PSUPP) 10-68
range and resolution 11-811-16
respiratory rate (f) 10-62
rise time % 10-69
TH:TL ratio 10-67
tidal volume 10-62
vent type 10-59
volume support (VT SUPP) 10-68
Ventilator Setup . . . . . . . . . . . . . . . 4-2
VentType . . . . . . . . . . . . . . . . . . . 10-59
Volume Support (VT SUPP) . . . . . . 10-68
VS
maximum pressure adjustments 10-26
startup 10-26
W
Warranty Information . . . . . . . . . 1-19
waveform axis scaling . . . . . . . . . 3-45

Part No. 10077893 A 2014-10


2014 Covidien.

Covidien llc
15 Hampshire Street, Mansfield, MA 02048 USA
,,.
www.covidien.com

[T] 1.800.255.6774

You might also like