Tbs 2013 May Weighing Instruments
Tbs 2013 May Weighing Instruments
Tbs 2013 May Weighing Instruments
BACK TO BASICS
A matter of balance:
calibration and use of
weighing instruments
In the modern laboratory, where commercially prepared consumables
now predominate, it is important to remember the importance of the
272
ANALYTICAL BALANCE
This type of instrument is
used to measure mass to
a very high degree of
MAY 2013
ARTICLE
BALANCE LOCATION
Balances must be sited on a firm, horizontal
surface. Many balances, especially those
measuring subgram quantities, will be fitted
with a type of spirit level and adjustable feet
to ensure a true level can be obtained.
A vibration-free balance stand not in contact
with both floor and wall may be required
when weighing extremely small quantities.
The environment must be free of draughts
and this is usually achieved by using an
instrument with a glass draught shield.
This is particularly important for balances
with 1 mg or 10 mg readability. Do not place
the balance next to a door, and avoid places
with high traffic. The effect of draught is
secondary for balances with readabilities
of 0.1 g.
Temperature fluctuations will also have
a detrimental effect on the weighing process
and most electronic balances now have
some form of automatic temperature
compensation.
Balances should also be protected against
magnetism and static electricity, both of
which have an adverse effect on the accuracy
of the balance. Most manufacturers offer
balances with antistatic weighing pans and
with draft shields coated with a substance
that conducts electricity.
Balance levelling and adjustment must
be performed to increase the sensitivity,
accuracy and efficiency of the device.
INSTRUMENT CALIBRATION
The standard for laboratory accreditation
is ISO 15189:2012, which specifies
requirements for quality and competence
in medical laboratories. At its most basic
level, calibration involves comparing a mass
reading against a standard weight. In the case
of a laboratory balance, standard weights are
graded according to class. There are four
ASTM weight classes (14), with Class 1
being the most accurate, and Class 4 being
MAY 2013
REPRODUCIBILITY
Reproducibility refers to the instruments
ability to deliver the same weight reading
repeatedly for a given object, and to return
to a zero reading after each weighing cycle.
This is usually stated as the standard
deviation that is determined from six to
10 measured values of a calibration weight.
LINEARITY
This test measures the ability of an
instrument to provide consistent sensitivity
throughout the weighing range. The test
requires several nominally equal weights,
each a fraction of the weighing capacity.
The group together should approximate
the weighing range of the instrument.
For example, a 160 g capacity analytical
balance might be tested with three 50 g
r
weights.
Other articles in this Back to Basics series
by Peter Riddle (prriddle@hotmail.co.uk)
are on centrifugation (February, page 76),
liquid-handling devices (March, page 138)
and pH meters and hydrogen ion
concentration (April, page 202).
273