Diamedix CH50
Diamedix CH50
Diamedix CH50
Sensitized Cells
SUMMARY OF PROCEDURE
1.
2.
3.
4.
5.
6.
INTENDED USE
For the determination of total complement activity (CH50) in human serum.
This assay is used to determine the functional integrity of the entire classic
complement pathway.
SPECIMEN COLLECTION
Whole blood should be collected by accepted medical techniques. A minimum
of 5 ml of whole blood is recommended. Allow blood to clot for approximately
60 minutes at room temperature (18-30 C). Centrifuge the sample and
transfer the serum to a clean tube at 2-8 C. Samples must be handled and
stored correctly to avoid erroneous results. If the serum is not tested on the
day it is separated, store at -70 C, preferably in aliquots. If storage does not
exceed 24 hours, serum can be kept frozen at -20 C. Prior to testing, bring
frozen sera to room temperature and mix gently avoiding foam formation. All
patient sera should then be kept cold (2-8 C) until used. Multiple freeze-thaw
cycles should be avoided. Hemolyzed specimens should not be used.
1. Calculation of Results
Results can be expressed either as % of the EZ Complement Reference
Serum or as CH50 values. Determine the results using the formula below:
Absorbance of Sample
Absorbance of Reference
a.
1.023
0.726
210
b.
1.023
0.726
105
= 148% of Reference
PROCEDURE
1.
2.
3.
4.
5.
6.
7.
8.
9.
10.
11.
Set up as many tubes as there are samples for testing plus one tube
each for the Reference and Low and High Controls and one tube for the
"spontaneous lysis" control. The "spontaneous lysis" control is a tube of
Sensitized Cells in buffer to which no sample is added.
Place the required number of tubes containing the Sensitized Cells in a
suitable rack and allow them to warm to room temperature (18 - 30 C)
for at least 60 minutes.
NOTE: In order to avoid incubation and mixing differences, Diamedix
recommends that runs be limited to 12-15 tubes.
VIGOROUSLY vortex or shake the tubes for 10 seconds to resuspend
the cells.
Remove the caps from all tubes. Add 5 l of patient samples, Reference
and controls to the appropriately labeled tube. After each sample
addition, replace the cap and mix IMMEDIATELY by shaking the tube
vigorously.
The "spontaneous lysis" tube should also be thoroughly mixed.
Allow the tubes to stand at room temperature (18 - 30 C) for 60 + 5
minutes.
Mix the contents of all tubes again by inverting 3-4 times.
Centrifuge the tubes at approximately 1800 RPM for 10 minutes.
Read the absorbances of the supernatants at 415 nm within 15 minutes
after centrifugation. Diamedix recommends the use of the EZ Reader for
absorbance determinations.
Read the absorbance of the "spontaneous lysis" control.
If the absorbance value is greater than 0.150, the results of the assay
are not considered valid. Repeat the test.
Zero the reader using the "spontaneous lysis" control as a blank. This
will correct for the degree of "spontaneous lysis" in the test samples.
Read and record the absorbance values of the Reference Serum and
each control and patient sample.
2. Interpretation of Results
% of Reference
0 to 50
51 to 150
> 151
2.
1.
2.
3.
EXPECTED VALUES
Normal values in populations can vary widely. Complement proteins are
acute-phase reactants and levels tend to increase with intercurrent illnesses.
There is also a tendency for levels to be higher in females who are either
pregnant or using oral contraceptives. In addition, it has been noted that
levels tend to rise slightly in aged individuals (4).
The EZ CH50 Test was performed using serum samples from one hundred
randomly selected, apparently healthy, blood donors from the S. Florida area.
These samples gave CH50 values ranging from 91 to 404. The mean CH50
value obtained was 274 with a Standard Deviation of 68. The distribution of
these values is shown in Figure 1.
FIGURE 1
Distribution of CH50 Values from 100 Normal Sera
35
30
25
Distribution
RESULTS
The concentration of Sensitized Cells has been adjusted to yield
approximately 50% hemolysis in the presence of 5 l of normal human serum.
Studies conducted by Diamedix have shown that the test is linear (R > 0.99)
for CH50 values of at least 400. Since the assay is linear over a broad range,
a single point can be used for calibration of this method.
Interpretation
Absent or low
Normal
High
LIMITATIONS
QUALITY CONTROL
1.
CH50 Value
0-100
101-300
> 301
20
15
10
5
0
50
100
150
200
250
300
350
400
450
CH50 Values
-2-
TABLE 1
EZ CH50 Results with Purchased Reference Materials
PERFORMANCE CHARACTERISTICS
A. Correlation Studies
The EZ CH50 Test was compared to another commercially available test for
the determination of total complement activity in human serum. Seventy-four
serum samples from normal blood donors were assayed by both methods. In
addition, a representative lot of EZ CH50 Reference, low and high controls
were tested by both methods. Also tested were several commercially
purchased deficient sera, Standard sera, as well as the WHO International
Reference Preparation (5) for a total of eighty-two samples. The correlation
between the methods is shown in Figure 2.
FIGURE 2
EZ CH50 Test vs. another commercially available test
CAE UNITS
250
Low
Normal
Sample
Factor B deficient
WHO Ref. Prep.
(Functional Whole
Complement)
Standard Complement
Serum
C3 deficient
Human Complement
Serum
EZ CH50 Value
3.0
100
94
122
N/A
112
3.0
N/A
221
High
200
D. Precision Testing
150
The precision of the test method was assessed by testing eight samples in
triplicate in two runs per day for three days. Samples included the in-house
'Gold' Standard, a representative EZ Reference Serum, EZ High Control
serum, EZ Low Control serum and four patient samples. Mean values are
expressed in CH50 units.
100
50
0
0
50
100
150
200
250
300
350
400
450
EZ CH50 VALUES
B. Linearity
An assessment was made of the linearity of the assay procedure. Different
amounts of sera were added to the Sensitized cell tubes and tested. Several
different serum samples were tested in this manner. These data indicated that
the assay is linear (R > 0.99) to at least 400 CH50 units. Figure 3 shows the
linearity of the in-house 'Gold' Standard material.
FIGURE 3
Linearity of EZ CH50
REFERENCES
1.
2.
3.
350
300
4.
250
CH50 Values
200
5.
150
100
50
0
2
C. Accuracy
The accuracy of the EZ CH50 Test was assessed by testing several
commercially available reference materials. The ability of the EZ CH50 test to
detect deficiencies in complement components was assessed by testing two
commercially available materials: one deficient for Factor B and the other
deficient for C3. Results are shown in Table 1.
I-789-001
Rev. 3 March 03
-3-
-4-