Operators Manual
FG2C/FG3C
Function Generator
Bedienungsanleitung
Manual de Instrucciones
Manuel dUtilisation
Bench Instruament
English O P E R A T O R S M A N U A L
E
N
Green/Yellow Earth
Blue
Neutral
Brown:
Live (Phase)
-1-
As the colours of the wires in main leads may not correspond with the colours marking identified in your plug/appliance, proceed
as follows:
The wire which is coloured Green & Yellow must be connected to the Earth terminal marked with the letter E or by the earth
symbol
or coloured Green or Green & Yellow.
The wire which is coloured Blue must be connected to the terminal which is marked with the letter N or coloured Blue or Black.
The wire which is coloured Brown must be connected to the terminal marked with the letter L or P or coloured Brown or Red.
If in doubt, consult the instructions provided with the equipment or contact the supplier. This cable/appliance should be
protected by a suitably rated and approved HBC mains fuse : refer to the rating information on the equipment and/or user
instructions for details. As a guide, cable of 0.75mm should be protected by a 3A or 5A fuse. Larger conductors would normally
require 13A types, depending on the connection method used.
Any moulded mains connector that requires removal /replacement must be destroyed by removal of any fuse & fuse carrier and
disposed of immediately, as a plug with bared wires is hazardous if engaged in a live socket. Any rewiring must be carried out in
accordance with the information detailed on this label.
CONTENTS
page
page
Warranty
Control Descriptions
5
5
Safety Information
Introduction
Specifications
MODEL
FG2C
FG3C
AM/FM modulation
na
SWEEP control
na
COUNTER input
na
GCV Output
na
TTL/CMOS output
VCF input
-2-
12a
1.5
1.2
1.8
100k
1M
10k
10
2.1
ATT
-20dB
2.4
0.9
2.7
0.6
3.0
0.3
CMOS
DUTY
OFFSET
OUTPUT
AMPL
50
FREQUENCY
POWER
5V
ADJ
DC 15V VCF
INPUT
MAX
100
1k
13
21
TTL/CMOS
OUTPUT
20
15V
CMOS ADJ
TTL
MIN
ADJ
MAX
-20dB
10
11
12
22
115
24
29W 36VA
AC
WARNING
TO AVOID SHOCK, REMOVE
INPUTS BEFORE OPENING.
FUSE RATING
115V
T0.315A 250V
230V
T0.16A 250V
IEC1010 250V CAT
FUSE
Fig. 2 FG2C Rear Panel
-3-
0.01
GATE
OVER
FREQUENCY
0.1
10
12a
17
Hz
1M
100k
10k
MOD
EXT
MOD
ON
GATE
MOD/DEPTH
SWEEP/TIME SWEEP/RATE
DUTY
100
1k
10
ATT
-20dB
CMOS
OFFSET
OUTPUT
AMPL
50
POWER
SLOW
13
FAST
LOG
LIN
SWEEP ON
SWEEP OFF
5V
MAX
AM
FM
15V
CMOS ADJ
ADJ
MIN
ADJ
MAX
-20dB
14
16
15
10
2a
11
12
22
24
115
18
COUNTER
INT
29W 36VA
EXT
AC
TTL/CMOS
OUTPUT
VCF/MOD
INPUT
GCV
OUTPUT
WARNING
INPUT
COUNTER
DC 15V
MAX
FUSE RATING
115V
T0.315A 250V
230V
T0.16A 250V
CAT
20
21
23
19
-4-
FUSE
CONTROL DESCRIPTIONS
FRONT PANEL
1.
Power Switch
Applies line power to the FG2C\FG3C power supply.
2.
4.
Counter Display
Shows the external frequency via a 6 digit green display, and the internal frequency via a 5 digit green display.
5.
Frequency Indicator
Indicates the current frequency value multiplier.
6.
7.
Table 1
Push button
Frequency Range
1
10
100
1k
10k
100k
1M
0.3Hz to 3Hz
3Hz to 30Hz
30Hz to 300Hz
300Hz to 3kHz
3kHz to 30kHz
30kHz to 300kHz
300kHz to 3MHz
8.
Function Selector
Press one of the three push buttons to select the desired output waveform.
9.
Duty Function
Pull out and rotate the knob to adjust the duty cycle of the waveform.
14 SWEEP/TIME
Control and LIN/LOG Selector:
(1) Rotate the knob clockwise to adjust sweep time for MAX, or counterclockwise for MIN.
(2) Select Linear sweep mode by pushing in the knob or LOG sweep mode by pulling out the knob.
15. MOD ON/off Selector
Press the button once, the indicator will light, and the output will be modulated by internal 400Hz sinewave or press the button
again, the indicator will be off, and the output will be modulated an external signal via VCF/MOD in connector (21) .
16. MOD/DEPTH
SWEEP RATE and AM/FM Selector:
(1) Sweep width can be controlled from 0 to 1000 times.
(2) Adjust modulation ratio by turning the knob clockwise for MAX, or counterclockwise for MIN.
(3) Select AM(amplitude modulation) mode by pushing in the knob or FM(frequency modulation) mode by pulling out the knob.
17 MOD EXT
Selector: Press the button once, the indicator will light, and EXTernal MODulation is selected. Press the button again, the
indicator will be off, and INT MODulation is selected.
FRONT or REAR PANEL
18. INT/EXT Counter Switch
Select internal counter mode (count the frequency of the FG3C output) or select EXT counter mode for an independent counter
(input signal from BNC (19) terminal).
19. EXT. Counter Input Terminal: Accepts external signals for measurement.
20. TTL/CMOS Output Terminal
TTL/CMOS compatible signal output
21. VCF/MOD
Input Terminal: Connector for the input voltage required to perform the "voltage control frequency" operation or the EXT
modulation operation.
22. Main Output Terminal
Main signal output.: The Output signal is calibrated for a 50 load.
23. GCV Output
This is a DC voltage output and its voltage level will follow the change of Frequency.
24. 115/230 Switch This switch selects the mains power voltage.
The functions of items 2, 2a, 3, 4, 5, 6, 14, 15, 16, 17, 18, 19, and 23 do not apply to model FG2C.
USAGE PROCEDURES
The following section describes the basic setup and controls for operating the functions of this instrument.
One of the best ways to observe waveforms is to connect the function generator to an Oscilloscope. Observe the effects that different
controls have on the waveforms displayed on the Oscilloscope when preceding with the following steps:
First-step check:
(1) Ensure that the mains voltage switch (24) on the rear panel of this instrument of is compatible with the mains supply.
Connect the instrument to the mains supply using the power cord supplied.
(2) Press PWR (1) switch, ensure all the rotary knobs are pushed in, and rotate the AMPL (12) knob to fully counterclockwise.
(3) Rotate the FREQ (13) control fully counterclockwise.
Triangle, square, and sine waves
(1) Select the Function (8) , select the Range (7), and rotate the FREQUENCY (13) knob to set the required frequency (read
output in display window (4) ).
(2) Connect the Output 22) terminal to the oscilloscope for observing the waveform signal.
(3) Rotate the AMPL (12) knob to control the waveform amplitude.
(4) If attenuation of the output signal is required, pull out the AMPL (12) knob to obtain 20dB attenuation or press the ATT -20dB
(12a) button for additional 20dB attenuation..
(5) The phase-relation of Output waveform and TTL output is shown in Figure 5 below:
-6-
Figure 5
TRIANGLE 0V
SINE 0V
SQUARE 0V
-7-
PRECAUTION ITEMS
(1) The DC OFFSET (11) knob, will provide a DC level voltage of 10V (no load) or 5V (50 load). However, the signal
amplitude plus the DC level, is limited to 20V (no load) or 10V (50 load). In case of over voltage, clipping will appear as
show n in Figure 6.
(2) The main OUTPUT terminal is calibrated into a 50 load. This indicates that the signal source impedance is 50. The OUTPUT
terminal may be connected to any circuit input impedance, but output the voltage and terminal impedance will not be calibrated.
To avoid oscillation, the OUTPUT terminal should be connected to a 50 load (especially when using high frequency and square
wave output), and the connecting cable should be as short as possible.
(3) When the DUTY knob is full counterclockwise, the ratio of positive state to negative state, should not be less than 80:20. When
the DUTY knob is full clockwise, the ratio of positive state to negative state, should not be greater than 20:80. Square waves can
be expanded to pulse waves, triangle waves can be expanded to ramp waves, sine waves can be expanded to unsymmetrical sine
waves. The Figure 7 shows the effect of the DUTY control adjustment on different waveforms.
Figure 6
Figure 7
VARY THE DUTY OF SHORTER TIME SECTION BY
ADJUSTING THE FREQUENCY ADJUSTMENT DIAL
A. Zero DC Offset
with Maximum Signal
PULSE
(SQUARE WAVE)
POSITIVE NEGATIVE
DC OFFSET DC OFFSET
B. Offset Limit
without Clipping
RAMP
(TRIANGLE WAVE)
POSITIVE NEGATIVE
DC OFFSET DC OFFSET
C. Excessive Offset
All examples:
Output terminated
is 50 ohm
UNSYMMETICRAL
(SINE WAVE)
TTL CMOS
POSITIVE NEGATIVE
DC OFFSET DC OFFSET
Warning: Failure to turn off and remove power from the Function Generator before changing the mains power
fuse could result in damage to the instrument and the operator.
Fuse replacement: Remove the fuse by inserting a small flat blade screwdriver into the notch at the bottom of the mains power
connector socket. Slide the fuse holder out. Replace the fuse with one of the following:
115 VAC 50/60 Hz: T0.315A/250V fuse (5mm x 20mm), I.R. 100kA Wavetek p/n 2400-05-0069, Littel Fuse p/n 218.315, or
Bussmann p/n. GDC-315mA.
230 VAC 50/60 Hz: T0.160A/250V fuse (5mm x 20mm), I.R. 100kA Wavetek p/n 2400-05-0070, Littel Fuse p/n 218.160, or
Bussmann p/n. GDC-160mA.
Warning: Use only an equivalent fuse to the one specified. Use of an incorrect fuse could result in serious injury or even death.
Reassembly: Slide the fuse holder back into the fuse cavity and reconnect the power cord.
MAINTENANCE & REPAIR
If there appears to be a malfunction during the operation of the Function Generator, the following steps should be performed in order
to isolate the cause of the problem: Review the operating instructions for possible mistakes in operating procedure. Inspect and
test the Test Cables for a broken or intermittent connection. Inspect and test the fuse. See Fuse Replacement. Except for the
replacement of the fuse, or test cables, repair of the Function Generator should be performed only by a Factory Authorized Service
Center or by other qualified instrument service personnel. The front panel and case can be cleaned with a mild solution of detergent
and water. Apply sparingly with a soft cloth and allow the function generator to dry completely before using. Do not use aromatic
hydrocarbons or chlorinated solvents for cleaning.
-8-
in U.S.A.
Wavetek Meterman
1420 75th Street SW
Everett, WA 98203
Tel: (877) 596-2680
Fax: (425) 446-6390
in Canada
Wavetek Meterman
400 Britannia Rd. E.Unit #1
Mississauga, ON L4Z 1X9
Tel: (905) 890-7600
Fax: (905) 890-6866
in Europe
Wavetek Meterman
52 Hurricane Way
Norwich, NR6 6JB, U.K.
Tel: int + 44-1603-404824
Fax: int + 44-1603-482409
SPECIFICATIONS
FG2C
FG3C
3Vpp
20 TTL load
25ns
3Vpp
20 TTL load
25ns
0 to 10V 1V(100:1)
10k 10%
0 to 10V 1V(100:1)
10k 10%
N/A
1.Main
Frequency Range
Amplitude
Impedance
Attenuator
DC Offset
Duty Cycle Control
LED Display
Range Accuracy
2.Sine Wave
Distortion
Flatness
3.Triangle Wave
Linear
4.Square Wave
Symmetry
Rise or Fall Time
5.CMOS Output
Level
Rise or Fall Time
6.TTL Output
Level
Fan Out
Rise or Fall Time
7.VCF
Input voltage
Input Impedance
8.GCV
Output voltage
-9-
FG2C
FG3C
N/A
N/A
N/A
N/A
Switch selector
100:1 ratio max. and adjustable
0.5 Sec to 30 Sec adjustable
Lin./Log. switch selector
N/A
N/A
N/A
N/A
0 to 100%
400 Hz(INT), DC to 1 MHz (EXT)
100 Hz to 3 MHz (-3dB)
10Vpp for 100 % modulation
N/A
N/A
N/A
0 to 5%
400Hz (INT), DC to 20 kHz (EXT)
10Vpp for 10% modulation
Int./Ext.
Range
Accuracy
Time base
Resolution
N/A
N/A
N/A
N/A
N/A
Input Impedance
Sensitivity
N/A
N/A
Switch selector
0.3 Hz to 3 MHz (5 Hz to 150 MHz EXT)
Time base accuracy 1count
20ppm (23C 5C) after 30 minutes warm up
The maximum resolution is 10 nHz for 1Hz
and 1 kHz for 100MHz
1M/150pf
35mVrms (5Hz to 100MHz)
45m Vrms (100MHz to150MHz)
9.Sweep Operation
Sweep/Manual
Sweep/Rate
Sweep/Time
Sweep/Mode
10.Amplitude Modulation
Depth
MODulation Frequency
Carrier BW
EXT Sensitivity
11.Frequency Modulation
Deviation
Modulation Frequency
EXT Sensitivity
12.Frequency Counter
13.General
LED Display
Power Source
Operation Environment
N/A
6 digits; 7.6mm (.3 in) high
115, 230VAC 15%, 50/60Hz
115, 230VAC 15%, 50/60Hz
Indoor use, Altitude up to 2000m. Ambient Temperature 0C to 40C
Relative Humidity 80%(maximum). EN61010-1, Installation category II, Pollution Degree 2
Storage Temperature & Humidity -10 C to 70C @ 70% R.H.(maximum).
Accessories
1 each RTL-1, Instruction Manual,
2 each RTL-1, Instruction Manual,
Spare Fuse
Spare Fuse
Dimension
243(w) x 93(H) x 292(D) mm
Weight
Approx. 2.0 kgs.
Approx. 2.2 kgs.
Safety:
Meets EN61010-1:1993. Installation category II Pollution degree 2
EMC: This product complies with requirements of the following European Community Directives: 89/336/EEC
(Electromagnetic Compatibility) and 73/23/EEC (Low Voltage) as amended by 93/68/EEC (CE Marking). However,
electrical noise or intense electromagnetic fields in the vicinity of the equipment may disturb the measurement
circuit. Measuring instruments will also respond to unwanted signals that may be present within the measurement
circuit. Users should exercise care and take appropriate precautions to avoid misleading results when making measurements in the
presence of electronic interference.
- 10 -
German
GEBRAUCHSANLEITUNG
Funktionsgeneratoren FG2C/FG3C
GEWHRLEISTUNG
Die Funktionsgeneratoren FG2C und FG3C sind ab Kaufdatum gegen Material- und Herstellungsfehler gewhrleistet fr die Dauer
von einem (1) Jahr. Ein Funktionsgenerator der whrend der Gewhrleistungsperiode reparaturbedrftig wird soll mit Kaufnachweis
zu einer Wevetek anerkannten Service Stelle oder zum Wavetek Hndler, wo das Gert gekauft wurde, gebracht/geschickt werden.
Siehe Abschnitt Wartung und Reparatur fr Einzelheiten. Irgendwelche implizierte Gewhrleistungen welche aus dem Kauf eines
Wavetek Funktionsgenerators fortflieen, einschlielich, jedoch nicht beschrnkt auf Verkuflichkeit oder Nutzbarheit zu einem
bestimmten Zweck, sind in der Dauer auf ein (1) Jahr beschrnkt. Wavetek ist nicht ansprechbar fr Gebrauchsverlu oder andere
Folgeschden und verbundene Kosten und Ausgaben. Verschiedene Staten und Lnder erlauben keine Beschrnkungen auf
Folgeschden oder deren Ausschlu. Deshalb sind diese Beschrnkungen vielleicht nicht fr Sie zupabar. Diese Gewhrleistung
gibt Ihnen bestimmte Rechte. Vielleicht haben Sie, je nach Staat oder Land, auch noch andere Rechte.
Zertifizierungen und Vorsichtsmanahmen
Dieses Gert ist EN61010-1:1993 zerifiziert fr Installationsklasse II, Pollutionsgrad 2. Alle Eingnge sind gegen dauerhafte
berbelastung geschtzt, bis zur Grenze welche je Funktion in den Spezifikationen angegeben oder auf dem Gert markiert ist (Siehe
Spezifikationen). berschreiten Sie niemals diese Grenzen. Prfen Sie Ihren Funktionsgenerator, Mekabel und Zubehr vor jeder
Messung. Im Problemfall (gebrochene oder beschdigte Mekabel, gerissenes Gehuse, keine Anzeige), Gert nicht verwenden.
Vermeiden Sie, beim Vornehmen einer Messung, jeden Kontakt mit Erdpotential. Berhren Sie keine Metallteile, Leiter, usw. Isolieren
Sie sich gegenber Erde (Kleidung, Schuhe, Teppich). Verwenden Sie das Gert nicht in einer explosiven Umgebung (entflammbare
Gase, Dampf, Staub). Verwenden Sie dieses, oder andere Megerte nicht ohne entsprechende Ausbildung.
GEFAHR Hohe Spannung
Schutzleiteransschlu
Gehuseanschlu
- 11 -
INHALT
Seite
Seite
Gewhrleistung
11
15
11
16
Sicherheitsinformation
11
18
Einleitung
12
Spezifikationen
19
GEBRAUCHSVORBEREITUNG AUSPACKEN
Die Verpackung sollte enthalten: ein Funktionsgenerator, ein Mekabel von 1 Meter Lnge (BNC nach Krokodilklemme), ein
Stromkabel, eine Ersatzsicherung und diese Anleitung. Sollte ein Teil fehlen oder beschdigt sein, bringen Sie das Gert zur
Kaufstelle zurck fr einen Umtausch.
EINLEITUNG
Die Funktionsgeneratoren FG2C und FG3C sind stabile Instrumente mit wenig Verzerrung, welche Sinus-, Dreieck- und
Rechteckwellenformen bis zu 3 MHz erzeugen. Das FG2C hat Kontrollen fr Amplitude, DC Komponente und Tastverhltnis.
Ausgnge sind Hauptausgang (50) und Logic-Ausgang (TTL oder CMOS). Das FG3C hat die gleichen Eigenschaften, plus interner
Wobbel (Log/Linear), Modulation (AM/FM), externer VCF/MOD, Zhlereingang und GCV (Spannung proportional zur Frequenz)
Ausgang. Der Zhler kann zum Zhlen einer externen Frequenz bis zu 150 MHz verwendet werden.
VERGLEICH DER FUNKTIONEN:
FUNKTION
FG2C
FG3C
AM/FM Modulation
MODELL
na
Wobbel Kontrolle
na
Zhlereingang
na
GCV Ausgang
na
TTL/CMOS Ausgang
VCF Eingang
Tastverhltnis Kontrolle
- 12 -
12a
1.5
1.8
1.2
100k
1M
10k
10
2.1
ATT
-20dB
2.4
0.9
2.7
0.6
3.0
0.3
CMOS
DUTY
OFFSET
OUTPUT
AMPL
50
FREQUENCY
POWER
5V
ADJ
DC 15V VCF
INPUT
MAX
100
1k
13
21
TTL/CMOS
OUTPUT
20
15V
CMOS ADJ
TTL
MIN
ADJ
MAX
-20dB
10
11
12
22
115
24
29W 36VA
AC
WARNING
TO AVOID SHOCK, REMOVE
INPUTS BEFORE OPENING.
FUSE RATING
115V
T0.315A 250V
230V
T0.16A 250V
IEC1010 250V CAT
SICHERUNG
Fig. 2 FG2C Rckseite
- 13 -
0.01
GATE
OVER
FREQUENCY
0.1
12a
17
10
Hz
1M
100k
10k
MOD
EXT
MOD
ON
GATE
MOD/DEPTH
SWEEP/TIME SWEEP/RATE
DUTY
100
1k
10
ATT
-20dB
CMOS
OFFSET
OUTPUT
AMPL
50
POWER
SLOW
13
FAST
LOG
LIN
SWEEP ON
SWEEP OFF
5V
MAX
AM
FM
15V
CMOS ADJ
ADJ
MIN
ADJ
MAX
-20dB
14
16
15
10
2a
11
12
22
24
115
18
COUNTER
INT
29W 36VA
EXT
AC
TTL/CMOS
OUTPUT
VCF/MOD
INPUT
GCV
OUTPUT
WARNING
INPUT
COUNTER
DC 15V
MAX
FUSE RATING
115V
T0.315A 250V
230V
T0.16A 250V
CAT
20
21
23
19
- 14 -
SICHERUNG
Ein/Aus-Schalter
Schaltet den FG2C/FG3C ein oder aus.
2.
Torzeit-Anzeige
Das Torintervall des internen Zhlers ist 0.01 Sekunde.
2a. Torzeitwahl
Drcken Sie diese Taste zum ndern der Torzeit in der EXTernen Zhlerbetriebsart. 0.01s, 0.1s, 1s, oder 10s stehen zur Wahl.
3.
berlastanzeige
Diese Anzeige leuchtet in der externen Zhler-Betriebsart wenn die Eingangsfrequenz den gewhlten Bereich berschreitet.
4.
Zhleranzeige
Zeigt die externe Frequenz auf einer 6-Digit Anzeige und die interne Frequenz auf einer 5-Digit Anzeige.
5.
Frequenzanzeige
Zeigt den Multiplikator der anstehenden Frequenz.
6.
Torzeitanzeige
Leuchtet kurz auf beim Beginn eines jeden Torintervalls.
7.
Tabelle 1
Taste
Frequenzbereich
Taste
Frequenzbereich
1
10
100
1k
10k
100k
1M
8.
Funktionswahl
Eine der drei Tasten drcken um die gewnschte Wellenform zu whlen.
9.
Taktverhltnis
Ziehen und drehen Sie diesen Knopf zum Einstellen des Taktverhltnisses.
Figur 5
Rechteck-Wellenform
TTL PULS 0V
DREIECK 0V
SINUS 0V
RECHTECK 0V
Dreieck-Wellenform
Taste, whlen Sie den Bereich mit Range (7), und drehen Sie den FREQUENCY (13)
(1) Drcken Sie die Function (8)
Knopf zum Einstellen der Frequenz.
(2) Verbinden Sie den Ausgang OUTPUT (22) mit dem Oszilloskop zur Betrachtung der Wellenform.
(3) Ziehen und drehen Sie den Duty (9) Knopf zum Einstellen der Impulsrampe.
(4) Drehen Sie den AMPL (12) Knopf zum Einstellen der Impulsamplitude.
(5) Ziehen Sie den AMPL (12) Knopf fr eine 20dB Abschwchung.
TTL/CMOS Ausgang
(1) Whlen Sie den Bereich mit Range (7), und drehen Sie den FREQUENCY (13) Knopf zum Einstellen der Frequenz.
(2) Verbinden Sei den TTL/CMOS (20) Ausgang mit dem Oszilloskop zum Betrachten der Wellenform.
(3) Der Ausgang ist ein TTL-Signal, fr allgemeine TTL integrierte Schaltungen verwendbar.
(4) Fr ein CMOS Signal, ziehen und drehen Sie den CMOS (10) Knopf um den gewnschten Signalpegel einzustellen.
Kontrolle der Frequenz durch externe Spannung (VCF)
Diese Betriebsart erlaubt die Einstellung der Frequenz mit einer extern angelegten DC Spannung.
(1) Whlen Sie die Wellenform mit Function (8) und den Bereich mit Range (7). Drehen Sie den FREQUENCY (13) Knopf um
die gewnschte Frequenz einzustellen.
(2) Legen Sie eine externe Spannung (010VDC) an den VCF (21) Eingang und beobachten Sie das Signal am Ausgang (22).
(3) Nehmen Sie andere Einstellungen vor mittels AMPL (12) Knopf zum Einstellen der Amplitude, schwchen Sie das Signal ab,
drehen Sie den Offset (11) Knopf zum Einstellen einer DC Komponente, drehen Sie den Duty (8) Knopf zum ndern des
Taktverhltnisses, usw.
Automatisches Wobbeln
(1) Whlen Sie die Wellenform mit einer Function (8) Taste und den Bereich mit Range (7). Drehen Sie den FREQUENCY (13)
Knopf zum Einstellen der Frequenz.
(2) Verbinden Sie den Ausgang (22) mit dem Oszilloskop zum Beobachten des Ausgangssignals.
(3) Drehen Sie den FREQUENCY (13) Knopf zum Einstellen der oberen Frequenzgrenze.
(4) Ziehen Sie den FREQUENCY (13) Knopf um die Wobbelfunktion zu aktivieren.
(5) Drehen Sie den SWEEP/TIME (14) Knopf und den SWEEP RATE (16) Knopf um die Wobbelzeit und Wiederholrate
einzustellen.
(6) Die LOG Wobbelbetriebsart wird gewhlt durch Ziehen des SWEEP/TIME (14) Knopfes.
Anmerkung: Die Wobbelbreite kann whrend einem Zyklus gendert aber der Wobbelzyklus kann nicht gestoppt werden.
AM/FM Betriebsart
(1) Whlen Sie die Wellenform mit einer Function (8) Taste und den Bereich mit Range (7). Drehen sie den FREQUENCY (13)
Knopf zum Einstellen der Frequenz.
(2) Verbinden Sie den Ausgang (22) mit dem Oszilloskop zum Beobachten des Ausgangssignals.
(3) Drcken Sie die MOD (15) Taste zur Wahl der FM Betriebsart und ziehen Sie den MOD/DEPTH (16) Knopf.
(4) Drcken Sie die MOD (15) Taste zur Wahl der AM Betriebsart und drcken Sie den MOD/DEPTH (16) Knopf.
(5) Drehen Sie den MOD/DEPTH (16) Knopf fr das gewnschte Modulationsverhltnis.
- 17 -
Vorsichtsmanahmen
(1) Der DC OFFSET (11) Knopf dient zum Einstellen eines DC Spannungspegels von 10V (ohne Last) oder 5V (mit 50 Last).
Die Signalamplitude, inklusiv DC Spannung, ist jedoch begrenzt auf 20V (ohne Last) oder 10V (mit 50 Last). Bei
berschrittener Grenze wird das Signal beschnitten, wie in Figur 6 gezeigt.
(2) Der Hauptausgang ist kalibriert fr eine 50 Last. Die Impedanz der Signalquelle ist ebenfalls 50. Der Ausgang kann mit
irgendwelcher Last verbunden werden, Ausgangsspannung und Impedanz sind dann jedoch nicht kalibriert. Um Oszillation zu
vermeiden, verbinden Sie den Ausgang mit einer 50 Last (speziell bei hoher Frequenz und Rechtecksignal). Das
Verbindungskabel sollte mglichst kurz sein.
(3) Wenn der DUTY Knopf ganz nach links gedreht ist, soll das Verhltnis zwischen positivem und negativem Teil nicht weniger als
80:20 sein. Wenn der DUTY Knopf ganz nach rechts gedreht ist, soll das Verhltnis zwischen positivem und negativem Teil nicht
hher als 20:80 sein. Rechtecksignale knnen in Impulssignale umgeformt werden, Dreiecksignal in Sgezahnsignale, und
Sinuswellenformen in unsymmetrische Sinussignale. The Figur 7 zeigt den Einfluss der DUTY Kontrolle auf verschiedene
Wellenformen.
Figur 6
Figur 7
IMPULSBREITE NDERN DURCH
FREQUENZREGELUNG
A. kein DC Offset
mit Max Signal
POSITIVER NEGATIVER
DC OFFSET DC OFFSET
IMPULS
(RECHTECK)
B. Offsetgrenze
ohne Beschnitt
RAMPE
(SGEZAHN)
POSITIVER NEGATIVER
DC OFFSET DC OFFSET
C. Zu viel Offset
Alle Beispiele:
Ausgang in
50 ohm
UNSYMMETISCHE
SINUSWELLE
TTL CMOS
POSITIVER NEGATIVER
DC OFFSET DC OFFSET
Warnung: Nicht-Abschalten des Gertes und Nicht-Entfernen des Netzkabels vor dem Sicherungswechsel kann
zu Schaden am Gert und Verletzungen fhren.
Sicherungswechsel: Alte Sicherung entfernen durch Einbringen der flachen Spitze eines kleinen Schraubenziehers in die ffnung
am Sockel des Netzspannungssteckers. Gleiten Sie den Sicherungshalter heraus. Ersetzen Sie die Sicherung wie folgt:
115 VAC 50/60 Hz: T0.315A/250V Sicherung (5mm x 20mm), I.R. 100kA Wavetek p/n 2400-05-0069, Littel Fuse p/n 218.315,
oder Bussmann p/n. GDC-315mA.
230 VAC 50/60 Hz: T0.160A/250V Sicherung (5mm x 20mm), I.R. 100kA Wavetek p/n 2400-05-0070, Littel Fuse p/n 218.160,
oder Bussmann p/n. GDC-160mA.
Warnung: Nur eine equivalente Sicherung verwenden. Sonst besteht Verletzungs- oder selbst Lebensgefahr.
Zusammensetzung: Sicherungshalter zurck in ffnung schieben und Netzkabel einstecken.
WARTUNG & REPARATUR
Sollte beim Anwenden des Gertes ein Problem auftreten, gehen Sie wie fogt vor um die Fehlerquelle zu bestimmen: Lesen Sie
die Anleitung um eventuelle Fehler bei der Bedienung festzustellen. Prfen Sie die Testkabel auf Bruch oder intermittente
Verbindung. Prfen Sie die Sicherung. Siehe Sicherungswechsel. Mit Ausnahme des Sicherungswechsels und Wechsels der
Testkabel soll eine Reparatur des Funktionsgenerators nur durch eine anerkannte Servicestelle oder durch Fachpersonal
vorgenommen werden. Das Gehuse kann mit einer milden Seifenlsung gereinigt werden. Sparsam mit einem weichen Tuch
auftragen und Gert vor erneutem Einsatz gut trocknen lassen.
- 18 -
in U.S.A.
Wavetek Meterman
1420 75th Street SW
Everett, WA 98203
Tel: (877) 596-2680
Fax: (425) 446-6390
in Canada
Wavetek Meterman
400 Britannia Rd. E.Unit #1
Mississauga, ON L4Z 1X9
Tel: (905) 890-7600
Fax: (905) 890-6866
in Europa
Wavetek Meterman
52 Hurricane Way
Norwich, NR6 6JB, U.K.
Tel: int + 44-1603-404824
Fax: int + 44-1603-482409
SPEZIFIKATIONEN
FG2C
FG3C
3Vss
20 TTL Last
25ns
3Vss
20 TTL Last
25ns
1. Hauptspezifikationen
Frequenzbereich
Amplitude
Impedanz
Abschwchung
DC Offset
Taktverhltnis
LED Anzeige
Bereichsgenauigkeit
2. Sinuswelle
Verzerrung
Flachheit
3. Dreieckwelle
Linearitt
4. Rechteckwelle
Symmetrie
Anstiegs-/Abfallzeit
5. CMOS Ausgang
Spannung
Anstiegs-/Abfallzeit
6. TTL Ausgang
Spannung
Belastbarkeit
Anstiegs-/Abfallzeit
N/A
FG2C
FG3C
N/A
N/A
N/A
N/A
Schalterwahl
100:1 max., einstellbar
0.5 Sek bis 30 Sek, einstellbar
Lin./Log. durch Schalterwahl
N/A
N/A
N/A
N/A
0 bis 100%
400 Hz (INT), DC bis 1 MHz (EXT)
100 Hz bis 3 MHz (-3dB)
10Vss fr 100 % Modulation
N/A
N/A
N/A
0 bis 5%
400Hz (INT), DC bis 20 kHz (EXT)
10Vss fr 10% Modulation
Int./Ext.
Bereich
Genauigkeit
Zeitbasis
Auflsung
N/A
N/A
N/A
N/A
N/A
Eingangsimpedanz
Empfindlichkeit
N/A
N/A
Schalterwahl
0.3 Hz bis 3 MHz (5 Hz bis 150 MHz EXT)
Zeitbasisgenauigkeit 1 Digit
20ppm (23C 5C) nach 30 Min Aufwrmung
Max. Auflsung ist 10 nHz fr 1Hz
und 1 kHz fr 100MHz
1M/150pF
35mVrms (5Hz bis 100MHz)
45mVrms (100MHz bis 150MHz)
9. Wobbelbetriebsart
Wobbel-/Normalbetrieb
Wobbel Verhltnis
Wobbel Zeit
Wobbel Modus
10.Amplitudenmodulation
Modulationstiefe
Modulationsfrequenz
Trgerwelle
EXT Empfindlichkeit
11.Frequenzmodulation
Abweichung
Modulationsfrequenz
EXT Empfindlichkeit
12.Frequenzzhler
13. Allgemein
LED Anzeige
Speisung
Betriebsumgebung
Lagertemperatur & Feuchte
Zubehr
Abmessungen
Gewicht
Sicherheit:
N/A
6 Digits; 7.6mm hoch
115, 230VAC 15%, 50/60Hz
115, 230VAC 15%, 50/60Hz
Innenhaus, Hhe bis 2000m. Umgebungstemperatur 0C bis 40C
Relative Feuchte 80%(max). EN61010-1, Installationskategorie II, Pollutionsgrad 2
-10 C bis 70C bei 70% R.F. (maximum).
1 RTL-1, Anleitung
2 RTL-1, Anleitung
Ersatzsicherung
Ersatzsicherung
243(B) x 93(H) x 292(T) mm
Approx. 2.0 kgs.
Approx. 2.2 kgs.
Gem EN61010-1:1993. Installationskategorie II, Pollutionsgrad 2
EMC: Dieses Produkt beantwortet an die Bestimmungen der folgenden EWG Richtlinien: 89/336/EEC
(Elektromagnetische Kompatibilitt) und 73/23/EEC (Niedrige Spannung) gendert durch 93/68/EEC (CE Marking).
Elektrisches Rauschen und starke magnetische Felder in der direkten Umgebung des Megertes knnen jedoch
den Mekreis beeinflussen. Das Gert kann auch durch Strsignale im gemessenen Schaltkreis beeinflut werden.
Der Anwender mu Vorsichtsmanahmen treffen um irrefhrende Meergebnisse bei Messungen in der Umgebung von starken
elektromagnetischen Feldern zu vermeiden.
- 20 -
Espaol M A N U A L D E L O P E R A D O R
- 21 -
CONTENIDO
pgina
pgina
Garanta
21
25
Certificaciones y precauciones
21
26
Informacin de seguridad
21
Mantenimiento y reparacin
28
Introduccin
22
Especificaciones
29
FG2C
FG3C
Modulacin AM/FM
MODELO
na
Control de barrido
na
Entrada de CONTADOR
na
Salida GCV
na
Salida TTL/CMOS
Entrada VCF
- 22 -
12a
1.5
1.8
1.2
100k
1M
10k
10
2.1
ATT
-20dB
2.4
0.9
2.7
0.6
3.0
0.3
CMOS
DUTY
OFFSET
OUTPUT
AMPL
50
FREQUENCY
POWER
5V
DC 15V VCF
INPUT
MAX
TTL/CMOS
OUTPUT
20
21
13
15V
CMOS ADJ
TTL
ADJ
100
1k
MIN
ADJ
10
MAX
-20dB
11
12
22
24
115
18
COUNTER
INT
29W 36VA
EXT
AC
TTL/CMOS
OUTPUT
VCF/MOD
INPUT
GCV
OUTPUT
WARNING
INPUT
COUNTER
DC 15V
MAX
FUSE RATING
115V
T0.315A 250V
230V
T0.16A 250V
CAT
20
21
23
19
- 23 -
FUSIBLE
0.01
GATE
OVER
0.1
FREQUENCY
12a
17
10
Hz
1M
100k
10k
MOD
EXT
MOD
ON
GATE
MOD/DEPTH
SWEEP/TIME SWEEP/RATE
DUTY
100
1k
10
ATT
-20dB
CMOS
OFFSET
OUTPUT
AMPL
50
POWER
SLOW
13
FAST
LOG
LIN
SWEEP ON
SWEEP OFF
5V
MAX
AM
FM
15V
CMOS ADJ
ADJ
MIN
ADJ
MAX
-20dB
14
16
15
10
2a
11
12
22
24
115
18
COUNTER
INT
29W 36VA
EXT
AC
TTL/CMOS
OUTPUT
VCF/MOD
INPUT
GCV
OUTPUT
WARNING
INPUT
COUNTER
DC 15V
MAX
FUSE RATING
115V
T0.315A 250V
230V
T0.16A 250V
CAT
20
21
23
19
FUSIBLE
Botn
Rango de frecuencias
Botn
Rango de frecuencias
1
0,3 Hz a 3 Hz
10 k
3 kHz a 30 kHz
10
3 Hz a 30 Hz
100 k
30 kHz a 300 kHz
100
30 Hz a 300 Hz
1M
300 kHz a 3 MHz
1k
300 Hz a 3 kHz
8. Selector de funciones
Presione uno de los tres botones para seleccionar la forma de onda de salida que desea.
9. Funcin de trabajo (DUTY)
Tire del mando y grelo para ajustar el ciclo de trabajo de la forma de onda.
10. Selector TTL/CMOS
Cuando el mando giratorio se encuentra presionado, del terminal BNC de salida TTL/CMOS OUTPUT (20) saldr una forma de
onda compatible TTL. Si el mando est hacia afuera, al girarlo ajustar la salida compatible CMOS (5 15 Vpp) de dicho
terminal.
11. Control de Offset CC (OFFSET)
Tire del mando giratorio para seleccionar cualquier nivel CC de la forma de onda entre 10 V; gire el mando a la derecha para
seleccionar un nivel positivo de Offset CC de la forma de onda, o hacia la izquierda para seleccionar un nivel negativo.
12. Control de amplitud de salida con funcin de atenuacin (AMPL)
Gire el mando a la derecha para obtener una salida mxima (MAX) y hacia la izquierda para obtener una salida mnima (MIN).
Tire del mando para obtener una atenuacin de salida adicional de 20 dB.
12a. Atenuacin de 20 dB (ATT 20dB)
Presione este botn para reducir la salida en 20 dB.
13 Selector de frecuencia (FREQUENCY) con Activacin y Desactivacin del barrido (SWEEP ON/OFF)
Gire el mando hacia la derecha para obtener frecuencias mximas y hacia la izquierda para frecuencias mnimas. (Mantenga el
puntero del mando dentro del rango de escalas del panel). Tire del mando giratorio para comenzar la operacin de barrido
automtico; el lmite superior de frecuencias viene determinado por la posicin del mando.
14 Control de tiempo de barrido (SWEEP/TIME) y Selector LIN/LOG:
(1) Gire el mando a la derecha para ajustar el tiempo de barrido al mximo (FAST), o hacia la izquierda para ajustarlo al mnimo
(SLOW).
(2) Para seleccionar el modo de barrido lineal (LIN), presione el mando giratorio; para seleccionar el modo de barrido LOG, tire
de l hacia afuera.
- 25 -
TRINGULO
0V
SENO 0 V
RECTNGULO
0V
Figura 5
- 27 -
Elementos de precaucin
(1) El mando giratorio OFFSET (11) ofrecer una tensin de nivel CC de 10 V (sin carga) o de 5 V (con carga de 50 ). No
obstante, la amplitud de la seal, ms el nivel de CC, est limitada a 20 V (sin carga) o 10 V (con carga de 50 ). En caso de
sobretensin, aparecer un truncamiento como muestra la figura 6:
(2) El terminal de salida principal OUTPUT est calibrado para una carga de 50 . Esto indica que la impedancia de la fuente de la
seal es 50 . El terminal OUTPUT pueden estar conectado a cualquier impedancia de entrada de circuito, pero no se calibrar
la tensin de salida ni la impedancia del terminal. Para evitar oscilaciones, el terminal OUTPUT deber conectarse a una carga
de 50 (especialmente cuando se trabaja con frecuencias altas y salidas de onda rectangulares) y el cable de conexin
deber ser lo ms corto posible.
(3) Cuando el mando giratorio DUTY est girado completamente a la izquierda, la proporcin de estado positivo a estado negativo
no deber ser menor que 80:20. Cuando el mando DUTY est girado completamente a la derecha, la proporcin de estado
positivo a estado negativo no deber ser mayor que 20:80. Las ondas rectangulares pueden ampliarse a ondas pulsatorias; las
ondas triangulares, a ondas con forma de dientes de sierra; y las ondas senoidales, a ondas senoidales asimtricas. La figura 7
muestra el efecto del ajuste del mando DUTY en diferentes formas de ondas.
Figura 6
Figura 7
A. Offset CC cero
con seal mxima
PULSO
(ONDA
RECTANGULAR)
OFFSET CC OFFSET CC
POSITIVA NEGATIVA
B. Lmite de Offset
sin truncamiento
DIENTES
DE SIERRA
(ONDA TRIANGULAR)
OFFSET CC OFFSET CC
POSITIVA NEGATIVA
ASIMTRICA
(ONDA SENOIDAL)
C. Offset excesivo
Todos los ejemplos:
Salida terminada
es 50 ohmios
OFFSET CC OFFSET CC
POSITIVA
TTL CMOS
NEGATIVA
en EE.UU..
Wavetek Meterman
1420 75th Street SW
Everett, WA 98203
Tel: (877) 596-2680
Fax: (425) 446-6390
en Canad
Wavetek Meterman
400 Britannia Rd. E.Unit #1
Mississauga, ON L4Z 1X9
Tel: (905) 890-7600
Fax: (905) 890-6866
en Europa
Wavetek Meterman
52 Hurricane Way
Norwich, NR6 6JB, U.K.
Tel: int + 44-1603-404824
Fax: int + 44-1603-482409
ESPECIFICACIONES
FG2C
FG3C
3 Vpp
20 carga TTL
25 ns
3 Vpp
20 carga TTL
25 ns
0 a 10 V 1 V (100:1)
10 k 10%
0 a 10 V 1 V (100:1)
10 k 10%
N/A
1. Principales
Rango de frecuencias
Amplitud
Impedancia
Atenuador
Offset CC
Control de ciclos de trabajo
Pantalla LED
Precisin de rango
2. Onda senoidal
Distorsin
Planidad
3. Onda triangular
Lineal
4. Onda rectangular
Simetra
Tiempo de subida o cada
5. Salida CMOS
Nivel
Tiempo de subida o cada
6. Salida TTL
Nivel
Salida en abanico
Tiempo de subida o cada
7. VCF
Tensin de entrada
Impedancia de entrada
8. GCV
Tensin de salida
- 29 -
FG2C
FG3C
N/A
N/A
N/A
N/A
Selector
100:1 de mxima y ajustable
0,5 a 30 s ajustable
selector Lin/Log
N/A
N/A
N/A
N/A
0 a 100%
400 Hz (INT), CC a 1 MHz (EXT)
100 Hz a 3 MHz (-3 dB)
10 Vpp para modulacin 100
9. Barrido
Barrido/Manual
Barrido/Tasa
Barrido/Tiempo
Barrido/Modo
10. Modulacin de amplitud
Profundidad
Frecuencia de modulacin
Ancho de banda de portadora
Sensibilidad EXT
N/A
N/A
N/A
0 a 5%
400 Hz (INT), CC a 20 kHz (EXT)
10 Vpp para modulacin 10%
Int./Ext.
Rango
Precisin
Base temporal
N/A
N/A
N/A
N/A
Resolucin
N/A
Impedancia de entrada
Sensibilidad
N/A
N/A
Selector
0,3Hz a 3MHz (5Hz a 150MHz EXT)
Precisin basada tiempo 1 unidad de cuenta
20 ppm (23 C 5 C) tras 30 min de
calentamiento
La resolucin mxima es 10 nHz para 1 Hz
y 1 kHz para 100 MHz
1 M/150 pf
35 m Vrms (5 Hz a 100 MHz)
45 m Vrms (100 MHz a150 MHz)
13. Generales
Pantalla LED
Fuente de alimentacin
Entorno de funcionamiento
N/A
6 dgitos; 7,6 mm (0,3 pulgadas) de altura
115, 230 V CA 15%, 50/60Hz
115, 230 V CA 15%, 50/60Hz
Uso en interiores. Altitud mxima: 2000 m. Temperatura ambiente: 0 a 40 C. Humedad
relativa 80% (mxima). EN61010-1, Categora de instalacin II, Grado de polucin 2
Temperatura y humedad de almacenamiento: -10 a +70C con una H.R. del 70% (mxima)
Accesorios
1 cada RTL-1, Manual de instrucciones.
2 cada RTL-1, Manual de instrucciones,
Fusible de repuesto
Fusible de repuesto
Dimensiones
243 (largo) x 93 (alto) x 292 (fondo) mm
Peso
2,0 kg, aprox.
2,2 kg, aprox.
Seguridad:
Cumple la norma EN61010-1:1993. Categora de instalacin II Grado de polucin 2.
EMC: Este producto cumple los requisitos de las siguientes Directivas de la Comunidad Europea: 89/336/EEC
(Compatibilidad electromagntica) y 73/23/EEC (Baja tensin) y de acuerdo con las modificaciones introducidas
por 93/68/EEC (Seales CE). En cualquier caso, la existencia de ruido elctrico o de campos electromagnticos
intensos cerca del equipo pueden afectar al circuito de medicin. Los instrumentos de medicin tambin
respondern a seales no deseadas que pueden estar presentes dentro del circuito de medicin. Los usuarios deberan tener cuidado
y tomar las precauciones necesarias para evitar resultados que puedan llevar a error al realizar las mediciones en presencia de
interferencias elctricas.
- 30 -
Franais M A N U E L
D U T I L I S AT I O N
- 31 -
TABLE DE MATIRES
Page
Page
Garantie
31
35
Certifications et prcautions
31
36
Information de scurit
31
Maintenance et rparation
39
Introduction
32
Spcifications
39
PRPARATIFS DBALLAGE
La boite demballage devrait contenir: le gnrateur de fonctions, un cble de 1m BNC vers pinces crocodile, un cble
dalimentation, un fusible de rserve, et ce manuel. Si une pice manque ou est endommage, retournez votre point de vente pour
un change.
INTRODUCTION
Les gnrateurs de fonctions FG2C et FG3C sont des instruments stables, faible distortion, qui gnrent des formes dondes
sinusodales, triangulaires et carres des frquences jusqu 3 MHz. Le FG2C possde des contrles damplitude, de dcalage et
de rapport de cycle. Les sorties sont Principale (Main) (50) et logique (TTL et CMOS). Le FG3C a les mmes fonctions que le
FG2C, plus balayage interne (Log/Linear), modulation (AM/FM), VCF/MOD externe, des entres de compteur et sortie GCV. Le
compteur peut tre utilis pour mesurer et afficher la frquence dun signal externe jusqu 150 MHz.
TABLE DE COMPARAISON DES MODLES:
CHARACTRISTIQUE / MODLE
FG2C
FG3C
Modulation AM/FM
na
Contrle de Balayage
na
Entre Compteur
na
Sortie GCV
na
Sortie TTL/CMOS
Entre VCF
- 32 -
12a
1.5
1.8
1.2
100k
1M
10k
10
2.1
ATT
-20dB
2.4
0.9
2.7
0.6
3.0
0.3
CMOS
DUTY
OFFSET
OUTPUT
AMPL
50
FREQUENCY
POWER
5V
ADJ
DC 15V VCF
INPUT
MAX
100
1k
13
21
TTL/CMOS
OUTPUT
20
15V
CMOS ADJ
TTL
10
MIN
ADJ
MAX
-20dB
11
12
22
115
24
29W 36VA
AC
WARNING
TO AVOID SHOCK, REMOVE
INPUTS BEFORE OPENING.
FUSE RATING
115V
T0.315A 250V
230V
T0.16A 250V
IEC1010 250V CAT
FUSIBLE
Fig. 2 Panneau arrire FG2C
- 33 -
0.01
GATE
OVER
0.1
FREQUENCY
12a
17
10
Hz
1M
100k
10k
MOD
EXT
MOD
ON
GATE
MOD/DEPTH
SWEEP/TIME SWEEP/RATE
DUTY
100
1k
10
ATT
-20dB
CMOS
OFFSET
OUTPUT
AMPL
50
POWER
SLOW
FAST
LOG
LIN
SWEEP ON
SWEEP OFF
5V
MAX
AM
FM
15V
CMOS ADJ
ADJ
MIN
ADJ
MAX
-20dB
13
14
16
15
10
2a
11
12
22
24
115
18
COUNTER
INT
29W 36VA
EXT
AC
TTL/CMOS
OUTPUT
VCF/MOD
INPUT
GCV
OUTPUT
WARNING
INPUT
COUNTER
DC 15V
MAX
FUSE RATING
115V
T0.315A 250V
230V
T0.16A 250V
CAT
20
21
23
19
FUSIBLE
Interrupteur
Met lappareil sous tension.
2.
Indicateur dIntervalle
Indicateur dintervalle de porte (lintervalle de porte du compteur interne est de 0.01 seconde).
4.
Afficheur
La frquence externe est affiche par un affichage vert de six digits, et la frquence interne par un affichage vert de cinq digits.
5.
Indicateur de Frquence
Indique le multiplicateur de la valeur de frquence en usage.
6.
7.
Tableau 1
Bouton
Gamme de Frequence
1
10
100
1k
10k
100k
1M
0.3Hz 3Hz
3Hz 30Hz
30Hz 300Hz
300Hz 3kHz
3kHz 30kHz
30kHz 300kHz
300kHz 3MHz
8.
Slecteur de Functions
Pressez lun des trois boutons pour slectionner la forme donde souhaite.
9.
Rapport de Cycle
Tirez et tournez ce bouton pour ajuster le rapport de cycle.
14 Temps de Balayage
Bouton de Contrle et LIN/LOG:
(1) Tournez le bouton dans le sens des aiguilles dune montre pour ajuster le temps de balayage vers MAX et dans le sens
contraire pour MIN.
(2) Slectionnez le mode Linaire en poussant le bouton ou le mode LOG en tirant le bouton.
15. Slecteur MOD actif/inactif
Pressez le bouton une fois le voyant sallume et la sortie est module par une onde sinusodale interne de 400Hz; pressez le
bouton nouveau le voyant steint et la sortie est module par un signal externe via VCF/MOD la borne (21) .
16. MOD/DEPTH (profondeur)
Slecteur SWEEP RATE (taux de balayage) et AM/FM:
(1) La plage de balayage peut tre contle de 0 1000.
(2) Adjustez le taux de modulation en tournant le bouton dans le sens des aiguilles dune montre pour MAX et dans le sens
contraire pour MIN.
(3) Slectionnez le mode AM (modulation amplitude) en poussant le bouton ou le mode FM (modulation de frquence) en tirant le
bouton.
17 Slecteur MOD EXT
Pressez le bouton une fois le voyant sallume et la MODulation EXTerne est slectionne. Pressez le bouton nouveau et la
MODulation INTerne est slectionne.
PANNEAU AVANT OU PANNEAU ARRIRE
18. Slecteur Compteur INT/EXT
Slectionnez le mode Compteur Interne (compte la frquence de la sortie du FG3C) ou compteur EXT pour compter la frquence
dun signal dentre la borne BNC (19).
19. Borne dentre EXT.: Sert connecter des signaux externes pour la mesure de la frquence.
20. Borne de sortie TTL/CMOS
Sortie de signal compatible TTL/CMOS
21. VCF/MOD
Borne dentre de tension pour le contrle de la frquence ou pour modulation EXT.
22. Borne de Sortie Principale
Le signal de sortie est calibr pour une charge de 50 .
23. Sortie GCV
La sortie est une tension CC, proportionnelle la frquence.
24. Slecteur 115/230
Permet dadapter lappareil la tension du secteur.
Les fonctions numrotes 2, 2a, 3, 4, 5, 6, 14, 15, 16, 17, 18, 19, et 23 ne sappliquent pas au modle FG2C.
PROCEDURES DUTILISATION
Cette section dcrit les procdures suivre pour utiliser les fonctions de lappareil.
La meilleure facon dobserver les formes donde, est de connecter le gnrateur de fonctions un oscilloscope. Ceci vous permet
dobserver les effets des diffrents contrles sur le forme donde affiche quand vous excutez les procdures dcrites ci-aprs.
Premier Contrle:
(1) Assurez-vous que le slecteur de tension (24) larrire de lappareil est bien mis sur la tension qui correspond la tension
du secteur. Raccordez lappareil au secteur en utilisant le cordon dalimentation fourni.
(2) Pressez le bouton PWR (1) , assurez-vous que tous les boutons rotatifs sont pousss, et tournez le bouton AMPL (12)
jusquau bout dans le sens contraire des aiguilles dune montre.
(3) Tournez le bouton FREQ (13) jusquau bout dans le sens contraire des aiguilles dune montre.
Ondes Triangulaire, Carre et Sinusodale
(1) Slectionnez Function (8) , slectionnez la gamme Range (7), et tournez le bouton FREQUENCY (13) pour ajuster la
frquence souhaite (suivez la lecture de lafficheur (4) ).
(2) Connectez la sortie Output (22) loscilloscope pour observer la forme donde.
(3) Tournez le bouton AMPL (12) pour contrler lamplitude du signal.
- 36 -
(4) Si une attnuation du signal de sortie est requise, tirez le bouton AMPL (12) pour obtenir un attnuation de 20dB, ou pressez le
bouton ATT -20dB (12a) pour une attnuation supplmentaire de 20dB.
(5) Le rapport de phase du signal de sortie et la forme du signal TTL sont montrs par la figure 5.
Gnration dUn Signal Carr
IMPULSION TTL 0V
TRIANGULAIRE 0V
SINUSOIDALE 0V
ONDE CARREE 0V
Figure 5
(3) Pour une modulation FM, pressez le bouton MOD (15) et tirez le bouton MOD/DEPTH (16).
(4) Pour une modulation AM, pressez le bouton MOD (15) et poussez le bouton MOD/DEPTH (16).
(5) Tournez le bouton MOD/DEPTH (16) pour ajuster le rapport de modulation souhait.
PRCAUTIONS
(1) Le bouton DC OFFSET (11) ajoute une niveau de tension CC de 10V (sans charge) ou 5V (avec une charge de 50).
Cependant, lamplitude du signal, y compris la composante CC est limite 20V (sans charge) ou 10V (charge de 50). En
cas de dpassement, le signal est tronqu, comme indiqu en figure 6.
(2) La sortie principale est calibre pour une charge de 50. Limpdance de source est galement de 50. La sortie du gnrateur
peut tre connecte un circuit avec nimporte quelle autre impdance dentre, mais la tension de sortie ainsi que limpdance
de sortie ne seront pas calibres. Pour viter un phnomne doscillation, la borne de sortie doit tre connecte une impdance
de 50 (spcialement dans le cas dune onde carre haute frquence), et le cble de connection doit tre le plus court
possible.
(3) Quand le bouton DUTY est tourn pleinement dans le sens contraire des aiguilles dune montre, le rapport de cycle positif par
rapport au cycle ngatif ne doit pas tre inrieur 80:20. Quand le bouton DUTY est tourn pleinement dans le sens des aiguilles
dune montre, le rapport entre le cycle positif et le cycle ngatif ne doit pas tre suprieur 20:80. Des ondes carres peuvent
tre modifies en ondes rectangulaires, des ondes triangulaires en dents de scie, et des ondes sinusodales en ondes
sinuodales dformes. La figure 7 montre leffet du rglage DUTY sur diffrentes formes donde.
Figure 6
Figure 7
CHANGEZ LE RAPPORT DU SEGMENT COURT
AVEC LE BOUTON DE CONTROLE DE FREQUENCE
A. Dcalage Zero
avec Signal Maximum
Dcalage
CC Positif
IMPULSION
(ONDE CARREE)
Dcalage
CC Ngatif
B. Limite de Dcalage
sanss Dcoupage
C. Dcalage Excessif
Tous les examples:
Sortie termine
par 50 ohm
DENTS DE SCIE
(ONDE TRIANGULAIRE)
Dcalage
CC Positif
Dcalage
CC Ngatif
Dcalage
CC Positif
Dcalage
CC Ngatif
NON-SYMETRIQUE
(SINUSOIDALE)
TTL CMOS
- 38 -
Attention: Ne pas teindre le gnrateur de fonctions, et ne pas retirer les cordons de mesure et dalimentation
avant de remplacer le fusible principal, peut endommager lappareil et blesser lutilisateur.
Remplacement du fusible: Retirez le fusible en insrant la lame dun petit tournevis plat dans la fente la base du connecteur
dalimentation. Glissez le porte-fusible vers lextrieur. Remplacez le fusible comme indiqu ci-dessous:
115 VAC 50/60 Hz: Fusible T0.315A/250V (5mm x 20mm), I.R. 100kA Wavetek p/n 2400-05-0069, Littel Fuse p/n 218.315, ou
Bussmann p/n. GDC-315mA.
230 VAC 50/60 Hz: Fusible T0.160A/250V (5mm x 20mm), I.R. 100kA Wavetek p/n 2400-05-0070, Littel Fuse p/n 218.160, ou
Bussmann p/n. GDC-160mA.
Attention: Utilisez seulement un fusible quivalent celui qui est spcifi. Lutilisation dun fusible incorrect peut entrainer des
blessures graves, voir fatales.
Rassemblage: Glissez le porte-fusible dans son logement et branchez le cordon damimentation.
MAINTENANCE & RPARATION
En cas de malfonctionnement pendant lutilisation du gnrateur, procdez comme suit pour isoler la cause du problme:
Consultez ce manuel pour voir sil ny a pas une erreur de procdure. Inspectez et testez les cordons de mesure pour dtecter une
ventuelle rupture ou un contact intermittant. Inspectez et testez le fusible. Voir Remplacement du Fusible. A lexception du
remplacement du fusible ou des cordons de mesure, une rparation du gnrateur de fonctions doit tre effectue par un centre de
services agre ou un technicien qualif.
Le panneau avant et le botier peuvent tre nettoys avec de leau et un dtergent doux. Appliquez parcimonieusement avec un chiffon
doux, et laissez scher lappareil compltement avant utilisation.
en U.S.A.
Wavetek Meterman
1420 75th Street SW
Everett, WA 98203
Tel: (877) 596-2680
Fax: (425) 446-6390
en Canada
Wavetek Meterman
400 Britannia Rd. E.Unit #1
Mississauga, ON L4Z 1X9
Tel: (905) 890-7600
Fax: (905) 890-6866
- 39 -
en Europa
Wavetek Meterman
52 Hurricane Way
Norwich, NR6 6JB, U.K.
Tel: int + 44-1603-404824
Fax: int + 44-1603-482409
SPCIFICATIONS
FG2C
FG3C
1.Spcifications Principales
Gamme de frquence
Amplitude
Impdance
Attnuateur
Dcalage CC
Contrle de Cycle
Affichage LED
Prcision de gamme
2. Onde Sinusodale
Distortion
Linarit
3. Onde Triangulaire
Plage
4. Onde Carre
Symmtrie
2%, 0.3 Hz 100 kHz
Temps de monte et de descente 100ns sortie maximum.
(dans 50)
5. Sortie CMOS
Niveau
Temps de monte/descente
3Vcc
20 charges TTL
25ns
3Vcc
20 charges TTL
25ns
0 10V 1V(100:1)
10k 10%
0 10V 1V(100:1)
10k 10%
6. Sortie TTL
Niveau
Puissance
Temps de monte/descente
7. VCF
Tension dentre
Impdance dentre
N/A
N/A
N/A
N/A
N/A
par slecteur
100:1 max., ajustable
0.5 Sec 30 Sec, ajustable
slecteur Lin./Log.
9. Balayage
Balayage/Manuel
Rapport de balayage
Temps de balayage
Mode de balayage
- 40 -
FG2C
FG3C
N/A
N/A
N/A
N/A
0 100%
400 Hz (INT), CC 1 MHz (EXT)
100 Hz 3 MHz (-3dB)
10Vcc pour modulation 100 %
N/A
N/A
N/A
0 5%
400Hz (INT), DC 20 kHz (EXT)
10Vcc pour une modulation de 10%
Int./Ext.
Gamme
Prcision
Base de temps
N/A
N/A
N/A
N/A
Rsolution
N/A
Impdance dentre
Sensibilit
N/A
N/A
Par slecteur
0.3 Hz 3 MHz (5 Hz 150 MHz EXT)
Prcison de la base de temps 1 digit
20ppm (23C 5C) aprs un chauffement
de 30 minutes
La rsolution maximum est 10 nHz pour 1Hz
et 1 kHz pour 100MHz
1M/150pF
35mVrms (5Hz 100MHz)
45m Vrms (100MHz 150MHz)
13.Gnral
Afficheur LED
Alimentation
Environnement dutilisation
N/A
6 digits; 7.6mm de hauteur
115, 230VAC 15%, 50/60Hz
115, 230VAC 15%, 50/60Hz
Utilisation intrieure, Altitude jusqu 2000m. Temprature ambiente 0C 40C
Humidit relative 80% (maximum). EN61010-1, Catgorie dinstallation II, Degr de pollution 2
Temprature stockage/humidit -10 C 70C, 70% H.R. (maximum).
Accessoires
Un RTL-1, Manuel,
2 RTL-1, Manuel,
Fusible de rserve
Fusible de rserve
Dimensions
243(L) x 93(H) x 292(P) mm
Poids
Approx. 2.0 kg
Approx. 2.2 kg
Scurit
Selon EN61010-1:1993. Catgorie dinstallation II, degr de pollution 2.
EMC: Ce produit est conforme aux exigences des directives suivantes de la Communaut Europenne: 89/336/EEC
(Compatibilit Electromagntique) et 73/23/EEC (Basse Tension), modifie par 93/68/EEC (CE Marking).
Cependant, du bruit lectrique ou des champs lectromagntiques intenses dans la proximit de linstrument peuvent
influencer le circuit de mesure. Linstrument peut galement tre perturb par des signaux parasytes dans le circuit
mesur. Lutilisateur doit tre vigilant et prendre des prcautions appropries pour viter des rsultats errons quand
les mesures sont prises en prsence dinterfrences lectromagntiques.
- 41 -