CG3 PDF
CG3 PDF
CG3 PDF
SCINTREX LIMITED
Head Office
In the U.S.
In Australia
Tel.:
(905) 669-2280
Fax:
(905) 669-6403
Telex: 06-964570
Tel.:
Fax:
Tel:
Fax:
(817) 591-7755
(817) 591-1968
(619) 321-6934
(619) 481-1201
Autograv CG3
Table of Contents
Section 1:
INSTRUMENT OVERVIEW
Section 2:
GETTING STARTED
Unpacking the Instrument . . . . . . . . . . . . . . . . . . . .21
Unpacking . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .22
Removing the Dust Cover . . . . . . . . . . . . . . . . . . . . .22
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.210
.210
.210
.211
.212
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.242
.243
.245
.246
.246
Table of Contents
Section 3:
INSTRUMENT PARAMETERS
Main Menu (Auxiliary Mode) . . . . . . . . . . . . . . . . . . 31
Auxiliary Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
INITIALISE Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
INITIALISE Submenus . . . . . . . . . . . . . . . . . . . . . . . . 33
OUTPUT . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 330
OUTPUT Submenus . . . . . . . . . . . . . . . . . . . . . . . . . 331
EDIT. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 339
LOCK AUX. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 340
Locking the Auxiliary mode . . . . . . . . . . . . . . . . . . . . 340
Unlocking Auxiliary mode . . . . . . . . . . . . . . . . . . . . . 340
Section 4:
FIELD OPERATION
Parameter Setup and Keypad Operation . . . . . . . . 41
Reading Time Terminating a Reading . . . . . . . . . . . 42
Cal. After . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
Levelling Requirements and
Continuous Tilt Correction . . . . . . . . . . . . . . . . . . . . 43
Recording Data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44
Auto Rejection . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44
Notebook Feature . . . . . . . . . . . . . . . . . . . . . . . . . . 45
..
..
..
..
..
.
.
.
.
.
..
..
..
..
..
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
. 47
. 48
. 49
. 49
. 410
Section 5:
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.52
.53
.54
.55
Charger . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .55
Charging Time . . . . . . . . . . . . . . . . . . . . . . . . . . . . .56
Charger Line Voltage . . . . . . . . . . . . . . . . . . . . . . . .56
Selecting an alternate voltage . . . . . . . . . . . . . .56
MAINTENANCE
Obtaining Accurate Drift Correction Results . . . . . . .62
Using a Chart Recorder. . . . . . . . . . . . . . . . . . .63
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.64
.64
.66
.68
.69
.610
.610
.611
.612
.612
.613
.614
.615
.616
.616
.617
Table of Contents
Section 7:
TROUBLESHOOTING
Section 8:
REFERENCE INFORMATION
Autograv Specifications . . . . . . . . . . . . . . . . . . . . . . 81
Standard Accessories . . . . . . . . . . . . . . . . . . . . . . . 82
Optional Accessories . . . . . . . . . . . . . . . . . . . . . . . . 83
Application Software . . . . . . . . . . . . . . . . . . . . . . . . 84
Training . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 84
Instrument Parts List . . . . . . . . . . . . . . . . . . . . . . . . 85
CG-3 Gravity Meter . . . . . . . . . . . . . . . . . . . . . . . . . 85
Options and Accessories . . . . . . . . . . . . . . . . . . . . . 85
Appendix A: THEORY OF
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
. 87
. 87
. 87
. 88
. 88
OPERATION
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A1
Control Console and Software . . . . . . . . . . . . . . . . . A5
Appendix B: ELASTIC
.
.
.
.
.
.
.
.
.
.
.
.
HYSTERESIS EFFECTS
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
. A12
. A13
. A15
. A16
. A16
. A17
AUTOGRAV
CONSTANTS SHEET
INDEX
Table of Contents
Autograv CG3/3M Manual v
Autograv CG3
Section
Introduction
INSTRUMENT OVERVIEW
Instrument Overview
Excellent protection from changes in ambient temperature and atmospheric pressure is achieved by sealing the Autograv sensing element in a temperature stabilized vacuum chamber. The wide operating temperature of -40C
to +45C (optional +35C or +55C) enables the operator to use the Autograv
in many environments. Since the sensor is made from non-magnetic fused
quartz, the Autograv is not affected by magnetic field variations (as long as
they are less than ten times the Earths magnetic field, i.e. 0.5mT).
Low drift is a result of the extremely stable operating environment of the
quartz elastic system. It allows the long term drift of the sensor to be accurately predicted and a real time software correction reduces it to less than
0.02 mGals per day.
Internal tilt sensors constantly supply the Autograv with information when
an operator selects the continuous tilt correction feature. If measurements
are taken on unstable ground, errors due to the movement of the Autograv
are automatically eliminated.
An automatic tidal correction can also be selected via the keyboard. The operator enters in the geographical location and time zone information and
the Autograv automatically calculates and applies a real time tidal correction to each reading.
The internal 12 volt rechargeable battery provides sufficient power to operate the Autograv throughout a normal survey day. An operator can check the
battery voltage at any time by pressing a key and viewing the display. When
the battery voltage is approaching the discharge level, the Autograv emits an
audible alarm. An optional cable enables the instrument to operate from
any external 12 volt DC power supply or battery which can supply 3 amperes
peak current. In the event of a main battery failure, a set of built-in miniature
batteries keeps the memory safe for several days.
Introduction
Autograv CG3 Manual 1-3
Instrument Overview
Figure 1
Autograv CG3
Section
GETTING STARTED
UNPACKING
THE INSTRUMENT
ATTENTION:
During shipment
the battery must be
disconnected.
If you have just
received your
CG3, the battery
is fully charged, but
disconnected.
Please reconnect
the battery (see
step 6 on page 2-5)
before operating
How To
Getting Started
Figure 2
UNPACKING
1. Pull up the tab of the link lock and turn the tab counter-clockwise to
transportation case to view the top of the Autograv dust cover plate.
5. Remove the Autograv from the transportation case and visually inspect
for any physical damage that may have occurred during transportation.
6. From the right hand side of the transportation case, remove the
Autograv accessories.
two latches and sliding the cover away from the instrument.
2. Place the dust cover back inside of the transportation case for safe
keeping.
Note: If you are shipping the Autograv, insert the dust cover to protect
the instrument.
STARTING
UP THE
AUTOGRAV
Starting-up the Autograv for the first time, or after it has been turned off for
more than 48 hours, requires the following steps and waiting periods.
1. Powering up the Autograv The charger powers up the
instrument and also charges the battery.
2. Warm-up period After you connect the battery charger to the
Autograv, it requires a 4 hour warm-up time in order to reach the
operating temperature.
You can unplug the battery charger from the Autograv to take test
readings during the stabilization period however, you must reconnect
the charger after the readings are complete or if the Autograv emits the
low battery alarm.
4. Checking and adjusting the temperature compensation
You check and adjust the temperature compensation after the
instrument is stable. (Refer to page 2-35)
5. Checking and adjusting the drift correctionYou check and
adjust the drift correction after the temperature compensation has
been checked (or adjusted if necessary). This procedure involves
running the Autograv for approximately 24 hours in the cycling mode
(unattended). (Refer to page 2-41)
6. Setting up the instrument for field operationsYou can set
up your Autograv for field use after you complete the previous steps.
(Refer to Section 4 - Field Operation)
How To
Getting Started
POWERING
UP THE
AUTOGRAV
1. Set the charger to the appropriate line voltage (115V or 230V) by adjust-
Fuse
115V
0.75A
230V
0.375A
3. After you have set the charger voltage and checked the fuse, plug the
power cord of the charger into an AC outlet. The green light on the
charger turns on to indicate that the charger is working.
Figure 3
2-4 Autograv CG3 Manual
pulling the black tab towards you so the battery can slide out of the
compartment.
8. Insert the connector of the battery into either socket on the left side of
the compartment. The Autograv has two sockets so that you can change
batteries without powering down the instrument (see Figure 4).
black tab against the bottom, side and top of the compartment. If the
black tab is not pressed against the walls of the compartment, the
battery will not fit and the foam padding will be damaged.
10.Place the battery inside of the compartment with the top of the battery
facing outwards and the battery cable to the left as shown in Figure 4.
11.Reinstall the battery compartment cover.
How To
9. Before you reinstall the battery, place your hand inside and flatten the
Getting Started
Figure 4
USING
THE
KEYPAD
How To
Figure 5
THE KEYPAD
You can use the Autograv keypad to:
Each key has two functions. The mode of operation determines which function the Autograv responds to. The two modes of operation are:
Operation modeIn this mode, the key press initiates the function that is
displayed at the top of each key.
Enter modeIn this mode, the key press initiates the function that is displayed at the bottom of each key.
Getting Started
Note: If the keypad is not responding to a key press, then you need
to reset the Autograv by pressing and holding the ON/OFF
(
) key for more than 5 seconds.
Description
ON/OFF Turns the instrument on and off.
r
1
2
38
4
memory by scrolling.
LINE Shows the present line number and/or a line
number recalled from memory (if applicable). The current line appears on the top line while the line number
from memory is on the second line.
It also enables you to change the current line number.
Key
5
0
6
Description
SCROLL KEYS Perform two functions:
1. Scroll through lists of selected data, actions or the
menu options
2. Goes to the next line or station if line or station
number is displayed
AUX Displays the AUXILIARY mode menu options.
7
9
How To
If you press the AUX key twice, then the software takes
you from your immediate menu to the first level menu
(i.e. the Main menu).
Getting Started
Key
Description
START/STOP Enables you to:
w
0
9
r
q
e
2-10 Autograv CG3 Manual
Description
CLEAR Erases the current number.
The Display
THE DISPLAY
PARAMETER SETUP DISPLAY
The Autograv display is an 80 character liquid crystal display (LCD) arranged in 4 rows of 20 digits in each row. For example:
DATA IDENTIFIER
100E
100N
0E
1
0
0
%
AVAILABLE
MEMORY
PRECEDING DATA
MENU STRUCTURE
The Autograv has three levels of menus. Refer to the Instrument Parameters
section (page 3-1) for a detailed description of the Autograv software.
How To
STATION
CHANGE?
LINE
STATION
PROMPT
CURRENT DATA
Getting Started
Example:
Press:
6A
UX
AUTOGRAV SETUP
ENTER?
1. By pressing ENTER, one item of the menu list appears on the display.
2. If you want to go to the next menu option in the same level, then:
Press:
5 or 0 (S
CROLL)keys
3. If you want to go to the submenu of the displayed menu, then press the
6 (A
UX)
key.
Selecting a Menu
When the display shows the menu option that you want to select, press the
key that is shown on the second line.
Example:
AUTOGRAV SETUP
ENTER?
KEY TO PRESS
Press:
e (E
NTER)
key.
Press:
e (C
HANGE)
key
The Display
Modifying Parameters
The prompt that is on the second line shows you how to change the parameter that is displayed. There are three types of prompts:
PROMPT
KEYS
SELECT/ENTER
xxxx> nnn
ENTER?
Example:
5 or 0
0 to 9,r,
w(C ),
q
LEAR
e (C
limited to selection
numerical values,
N,S,E,W, +/
limited to 2 selections
HANGE)
5 0
READ TIME
CHANGE?
Press:
e (C
65
HANGE)
READ TIME>65
ENTER
How To
xxxx
CHG? >xxxx<
TYPE
Getting Started
Type:
120(120)
READ TIME> 120
ENTER
Press:
e (E
NTER)
READ TIME
CHANGE?
120
The Display
How To
Autograv CG3 Manual 2-15
Getting Started
The Display
Figure 6
How To
Autograv CG3 Manual 2-17
Getting Started
Display
66
(A ) twice
Current Menu
UX
5 or 0 until...
e
Current Menu
5 or 0 until...
e(C
INITIALISE
ENTER?
HANGE)
TIME
CHANGE?
xx:xx:xx
xx/xx/xx
TIME
ENTER
HH:xx:xx
Display
HH
(Value for hour)
TIME
ENTER
HH:xx:xx
NTER)
TIME
ENTER
xx:MM:xx
MM
(Value for minutes)
TIME
ENTER
xx:MM:xx
NTER)
TIME
ENTER
xx:xx:SS
SS
(Value for seconds)
TIME:
ENTER
xx:xx:SS
e(E
DATE:
ENTER
xx/xx/xx
e(E
e(E
NTER)
How To
Key to Press
Getting Started
2. To Erase Memory
Key to Press
Display
66
Current Menu
(AUX) twice
5 or 0 until...
e(E
NTER)
5 or 0 until...
e(E
NTER)
q(S
TART/STOP)
INITIALISE
ENTER?
Current Menu
ERASE MEMORY
ENTER?
ERASE MEMORY
START?
MEMORYFREE 100%
TESTED XXK
XXK is the size of the memory installed in the Autograv. For example: 48K
3. Autograv Setup
Key to Press
66
(AUX twice)
Cycling and
Read Time
Setup
5 or 0 until...
NTER
e (C
HANGE)
5 or 0 until...
e (E
NTER)
(SCROLL down)
e (C
HANGE)
120
(120)
Current Menu
AUTOGRAV SETUP
ENTER?
Current Menu
MODE:XXXX
CHANGE?
MODE:XXXX
SELECT/ENTER
MODE:>CYCLING
SELECT/ENTER
MODE:CYCLING
CHANGE?
READ TIME XXX
CHANGE?
READ TIME>XXX
ENTER
READ TIME>120
ENTER
How To
e (E )
5 or 0 until...
Display
Getting Started
Key to Press
e(E
NTER)
(SCROLL down)
e (C
HANGE)
600
(600)
Chart Select
Setup
e (E
NTER)
(SCROLL down)
e (C
HANGE)
555 until...
STOP ERR
Setup
e (E
NTER)
(SCROLL down)
Display
Key to Press
ee
(C
until...)
HANGE
(SCROLL down)
e (C
HANGE)
(12)
CONT TILT
Setup
e (E
NTER)
0
Auto
Rejection
Setup
CAL. AFTER XX
CHANGE?
CAL. AFTER>12
ENTER
CAL. AFTER>12
ENTER
CAL. AFTER 12
CHANGE?
(SCROLL down)
CONT.TILT CORR.
CHG?>XXXX<
ee
(C
until...)
CONT.TILT CORR.
CHG?>DISABLED<
HANGE
0
TIDE CORR
Setup
STOP:ERR.<LIM
CHG?>DISABLED<
(SCROLL down)
AUTO REJECTION
CHG?>XXXX<
ee
(C
until...)
AUTO REJECTION
CHG?>SELECTED<
HANGE
(SCROLL down)
TIDE CORRECTION
CHG?>XXXX<
How To
12
Display
Getting Started
Key to Press
LAT and LONG
Setup
ee
(C
until...)
HANGE
(SCROLL down)
e (C
HANGE)
Decimal degrees of
the reading site
Display
TIDE CORRECTION
CHG?>SELECTED<
DEG. LAT XX
CHANGE?
DEG. LAT XX
ENTER
DEG. LAT>DD
ENTER
N is + (positive)
S is (negative)
e (E
NTER)
(SCROLL down)
e (C
HANGE)
Decimal degrees of
the reading site
DEG. LAT DD
CHANGE?
DEG. LONG XX
CHANGE?
DEG. LONG XX
ENTER
DEG. LONG>DD
ENTER
W is + (positive)
E is (negative)
e (E
NTER)
DEG. LONG DD
CHANGE?
GMT Setup
Key to Press
(SCROLL down)
e (C
HANGE)
(GMTClock time)
in hours
Display
GMT. DIFF XX
CHANGE?
GMT. DIFF>XX
ENTER
GMT. DIFF>HH
ENTER
AUTO
RECORD
Setup
e (E
NTER)
0
GREF Setup
GMT. DIFF HH
CHANGE?
(SCROLL down)
AUTO RECORD
CHG?>XXX<
ee
(C
until...)
AUTO RECORD
CHG?>DISABLED<
HANGE
(SCROLL down)
e (E
NTER)
0 (0 )
e (E
NTER)
GREF. XX
CHANGE?
GREF.>XX
ENTER
GREF.>0
ENTER
GREF. 0
CHANGE?
Autograv CG3 Manual 2-25
How To
Getting Started
DRIFT CORR
Setup
Key to Press
(SCROLL down)
e (E )
5 or 0 until...
NTER
e (C
HANGE)
Numerical value
(From constant
sheet or most recent
value)
e (E
NTER)
Display
DRIFT CORRECT.
ENTER?
Current Menu
DRIFT XX
CHANGE?
DRIFT>XX
ENTER
DRIFT>VV
ENTER
DRIFT VV
CHANGE?
(SCROLL down)
ee) twice
(C
HANGE
The drift correction start time is now set to the current time and date on
the clock.
6 (A
UX)
DRIFT CORRECT.
ENTER?
Initial
Constants
Setup
Key to Press
Display
(SCROLL down)
INIT. CONST.
ENTER?
(E
) and
e
then 5 or 0
) until...
(S
GCAL.1 XX
CHANGE?
NTER
CROLL
e (C
HANGE)
GCAL.>XX
ENTER
(Every instrument has its own set of initial constants. These must match
the parameters setup in the INITIAL CONSTANTS mode.)
e (E
NTER)
(SCROLL down)
GCAL. 1 XX
CHANGE?
GCAL. 2 XX
CHANGE?
Follow the key press steps (as shown in the previous four steps) for all five
of the initial constants and set the values to match the instruments individual constants. sheet values.
6 (A
0
UX)
(SCROLL down)
INIT. CONST.
ENTER?
MODE: CYCLING
CHANGE?
You are now back to the top/start of the AUTOGRAV SETUP submenu.
How To
Getting Started
THE DISPLAY
DURING
READINGS
After you press the START key to initiate a reading, the display goes through
two display changes to enable you to level the Autograv before you actually
start a reading.
The first display indicates that the Autograv is in the ADJUST mode. Within a
few seconds, the display changes to show you the values of the X and Y tilts,
as follows:
LEVELING PARAMETERS
TILTS
ADJUST
LINE
STATION
Y:
VALUES
X:XXX 1
XXX
0
100
0
0
%
AVAILABLE
MEMORY
READ
VERTICALLY
To level the instrument, read the display and analog tilt meters, and then adjust the tripod until the Autograv is level.
GRAV.
DUR
LINE
STATION
GRAVITY READING
XX.XXX
1
XER. X.XXX0
100
0
0
%
AVAILABLE
MEMORY
READ
VERTICALLY
How To
Getting Started
SETTING
UP THE
TRIPOD
The tripod enables you to level the Autograv before taking a reading. The
three leveling screws on the bottom of the tripod are adjusted until the X and
Y meters on the front panel of the Autograv and the display show that the
Autograv is level.
Figure 7
1. Set the tripod on stable ground in a quiet location with the bubble level
How To
AUTOGRAV
ADJUST
TILTS
ADJUST
X: XXX
Y: XXX
6. View the X and Y meters on the front panel and the display to level the
instrument.
Getting Started
Figure 8
To Level
You have two options for leveling the Y-axis of the CG-3/3M. The fastest way
to level the meter is to adjust both rear screws on the tripod as outlined in
Step 7. The other option for high precision microgravity measurements is to
set one of the rear screws in the middle of its range and tape over it. Adjust
the other rear screw until the Y-axis is level. This method ensures that the elevation of the CG-3/3M sensor relative to the tripod is strictly maintained.
The tripod comes with three extender legs which may be used under soft
soil conditions.
To set up
1. Place one extender leg over one of the pointed bottom legs of the tri-
pod with the tab of the extender leg facing the outside of the tripod.
This tab is available so that you can step on the tab to dig the shaft of
the leg into the soil.
2. Insert the 8-32 hex bolt through the openings of the legs. Ensure that the
head of the nut fits into the recessed slot on the extender leg as this
enables you to securely tighten the bolt.
3. Place the washer and the black wing nut onto the bolt and slightly
How To
Getting Started
Figure 9
Temperature Compensation
TEMPERATURE COMPENSATION
Important: Do not adjust the temperature compensation until the
Autograv has gone through the complete stabilization
time of 48 hours. If you can not wait for 48 hours, then
please refer to Using the CG-3 before Warm-up Stabilization is Complete on page 4-10.
CHECKING
1. Select a stable base for the Autograv in a quiet location and place the
TILTS
X: XXX
ADJUST Y: XXX
press the
0 (S
CROLL DOWN)
TEMP. X.XX
ADJUST
How To
Getting Started
SCREWS TO BE
REMOVED
Temperature Compensation
How To
Autograv CG3 Manual 2-37
Getting Started
Temperature Compensation
To Adjust
2.
Using a slot head screwdriver, remove the four M2.5 x 5mm flat
head screws from the Temperature Coarse Adjustment cover.
3.
4.
View the position of the Coarse Adjustment screw. You only need
to slightly move the screw to make an adjustment.
BEFORE TURNING THE SCREW, CAREFULLY NOTE ITS POSITION.
Clockwise = positive
Counter-clockwise = negative
After making a very small adjustment of a few degrees, wait
several seconds for the system to stabilize before reading the
temperature value.
Note: Do not push down on the coarse adjust screw, as this could
damage the adjustment potentiometer.
5.
6.
7.
How To
1.
Getting Started
8.
Note: If the reading does not come within the range with a coarse adjustment of 1 turn in either direction, contact Scintrex Limited.
9.
1. Measure Drift
1. After you adjust the temperature compensation, press the
(SCROLL DOWN) key to display the X and Y Tilts.
2. Level the instrument (using the tripod setup procedures) to within -10
How To
Getting Started
2. Output Data
6
50
6 (A ) key
2. Press the 5 or 0 (S
appears:
1. Press the
UX
CROLL)
OUTPUT
ENTER?
Note: You must setup the baud rate, data bits and carriage return of
the printer before you dump data. Refer to the Instrument
Parameter Section, starting on page 3-37, on how to set these
parameters.
AUTOGRAV
START DUMP
q
Example
DRIFT
R1,T1
GREF.:
-4158. mGals
Tilt x sensit.:
296.5
GCAL.1:
6131.436
Tilt y sensit.:
283.8
GCAL.2:
0.
Deg.Latitude:
43.7
TEMPCO.:
-0.1241 mGal/mK
Deg.Longitude:
79.6
Drift const.:
0.18
GMT Difference:
4.hr
Drift Correction Start Time: 11:57:16
Cal.after x samples:
12
Date: 95/04/13
On-Line Tilt Corrected = "*"
------------------------------------------------------------------------------Station Grav.
SD.
Tilt x Tilt y
Temp.
E.T.C. Dur # Rej
Time
0.
41.684 0.041
2.
-2.
-0.69
-0.023
120
0 14:03:57
0.
41.685 0.037
1.
-3.
-0.67
-0.023
120
0 14:06:27
0.
41.684 0.044
2.
-3.
-0.65
-0.023
120
1 14:08:57
0.031
0.036
0.034
0.033
0.045
0.036
0.038
0.038
0.042
0.034
0.043
0.035
0.037
0.037
0.032
0.039
0.039
0.049
0.039
0.038
0.040
0.038
0.042
0.044
0.040
0.038
0.039
0.036
0.043
0.041
0.044
0.040
1.
1.
1.
2.
2.
2.
1.
1.
1.
1.
2.
2.
2.
2.
2.
2.
2.
2.
2.
2.
2.
2.
2.
2.
2.
2.
2.
2.
2.
2.
2.
2.
-3.
-3.
-3.
-3.
-3.
-3.
-3.
-3.
-4.
-4.
-3.
-3.
-3.
-3.
-3.
-3.
-3.
-3.
-3.
-3.
-3.
-3.
-3.
-3.
-3.
-3.
-3.
-3.
-3.
-3.
-3.
-3.
0.
0.
0.
0.
0.
0.
0.
41.749
41.742
41.743
41.741
41.749
41.749
41.745
0.039
0.037
0.044
0.041
0.041
0.040
0.031
1.
2.
2.
1.
1.
1.
1.
-3.
-3.
-3.
-3.
-3.
-3.
-3.
-0.65
-0.65
-0.64
-0.64
-0.64
-0.64
-0.64
-0.64
-0.63
-0.64
-0.64
-0.64
-0.64
-0.63
-0.63
-0.63
-0.64
-0.64
-0.64
-0.64
-0.63
-0.64
-0.63
-0.63
-0.63
-0.63
-0.63
-0.63
-0.63
-0.63
-0.63
-0.63
.
.
.
.
.
-0.63
-0.63
-0.63
-0.63
-0.63
-0.63
-0.63
-0.023
-0.024
-0.024
-0.024
-0.024
-0.024
-0.023
-0.023
-0.023
-0.023
-0.023
-0.023
-0.023
-0.023
-0.023
-0.023
-0.023
-0.022
-0.022
-0.022
-0.022
-0.022
-0.022
-0.021
-0.021
-0.021
-0.021
-0.020
-0.020
-0.020
-0.020
-0.019
120
120
120
120
120
120
120
120
120
120
120
120
120
120
120
120
120
120
120
120
120
120
120
120
120
120
120
120
120
120
120
120
0
0
0
0
0
1
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
14:11:27
14:13:57
14:16:27
14:18:57
14:21:27
14:23:57
14:26:27
14:28:57
14:31:27
14:33:57
14:36:27
14:38:57
14:41:27
14:43:57
14:46:27
14:48:57
14:51:27
14:53:57
14:56:27
14:58:57
15:01:27
15:03:57
15:06:27
15:08:57
15:11:27
15:13:57
15:16:27
15:18:57
15:21:27
15:23:57
15:26:27
15:28:57
-0.016
-0.016
-0.016
-0.015
-0.015
-0.014
-0.014
120
120
120
120
120
120
120
0
0
0
1
0
0
0
01:51:27
01:53:57
01:56:27
01:58:57
02:01:27
02:03:57
02:06:27
the recording period and another reading (R2) at time T2 from near
the end of the period.
Do not use readings which are obviously noisy or disturbed.
2. From these values, calculate a new Drift Correction.
R2 R1
DRIFT = DRIFT + --------------------T2 T1
How To
R2,T2
0.
0.
0.
0.
0.
0.
0.
0.
0.
0.
0.
0.
0.
0.
0.
0.
0.
0.
0.
0.
0.
0.
0.
0.
0.
0.
0.
0.
0.
0.
0.
0.
Getting Started
where:
DRIFT is the Drift constant in the instrument during readings
(T2-T1) is in units of days.
Example:
The data from the printout on page 2-42 was used to calculate the following
drift value.
Selecting:
T2 = 02:03:57 on 95/05/09
T1 = 14:03:57 on 95/05/08
and
R2 R1 = 0.065mGal
T2 T1 = 12 hrs = 0.50 days
Therefore,
0.065
DRIFT = 0.18 + --------------0.5
5 or 0 (S
AUTOGRAV
SETUP ENTER?
CROLL)
key until
50
3. Press the
appears:
5 or 0 (S
CROLL)
DRIFT CORRECT.
ENTER?
e
50
appears:
5 or 0 (S
CROLL)
With a new instrument, check the Drift Correction once a week for 4 weeks.
After this period, check approximately once a month. For most sensors, the
drift rate will go down as they age, especially during the first couple of
months. Please also note that the drift correction start time is automatically
reset to zero after 128 days since the last adjustment giving rise to an apparent large offset. Therefore, you must reset the drift correction start time
before 128 days have elapsed.
How To
DRIFT XX
CHANGE?
Getting Started
MODE: CYCLING
CHANGE?
e
0
6. Take a reading by pressing the START key once to get into the ADJUST
mode.
7. Re-level the Autograv.
Recalling Data
RECALLING DATA
3 8
In memory, you can scroll from station to station, from line to line or from
grid to grid. The two MEMORY keys enable you to scroll sequentially through
the stations on a line, through the lines of the survey or through the grids of
the survey.
Various data recorded from one station can also be examined. The two
scroll keys enable you to scroll through the data or notebook information at
a particular station.
Note: This information is for recalling data in the FIELD mode for the
day on which it was recorded.
To recall data from previous days or to recall data in the
CYCLING mode, refer to pages 2-48 to 2-51.
How To
SCROLLING IN MEMORY
Getting Started
2D
9S
4L
7I
ATA
TATION
INE
NFO
4. Press one of the memory keys to enter the MEMORY mode. The upper
line of the display shows the current information, while the second line
shows the information stored in the memory. The prompt MEMORY
indicates the stored information.
Notes:
5 0
7
3
8
2
Recalling Data
3
8
The following signs appear on the second line to show you where
you are in the memory:
Sign
The message NO STN or NOLINE (or a similar message) indicates that there is no data for the station, line or grid number with
the number larger (
Memory) or smaller (
Memory)
than currently being examined.
)M
38
Use the ( 50) S
record.
5. Use the (
How To
Getting Started
9+ 38 1. Press the S
TATION
Press: CHANGE/ENTER
Enter: Station number and Direction
Press: CHANGE/ENTER
Note: The entry appears on the second line.
3. If the station is on a different line, then do the following:
4
e
e
2
Press: CHANGE/ENTER
4. Press: DATA key
5. Examine the data for the record.
Recalling Data
2
3OR 8
5OR 0
3 8
Press: 5or 0 to scroll through the parameters of a particular
or
MEMORY key to scroll through records (stored
Press:
in the same order in which they were made)
record.
4
8
2
or,
How To
Getting Started
Autograv CG3
Section
INSTRUMENT PARAMETERS
The Autograv operation is controlled through the keyboard accessed firmware functions. The software programs are permanently stored in EPROM
(erasable programmable memory) chips and are called firmware. There are
two sets of EPROM chips:
The firmware version of each set is indicated on the first line of the data
dump (page 331).
ENU
UX
3.
4.
6A
UX key.
Please refer to the software flowchart (Figure 6 on page 2-17) to view all of
the menu levels. A description of each parameter is given in the pages following the AUXILIARY mode software flowchart.
Parameters
The Autograv software is accessed using the keys on the keypad. The
AUXILIARY mode displays the main menu of the Autograv.
Instrument Parameters
Notes:
Auto-shutoff If you do not use the Autograv for one minute, the display
automatically turns off. Press the
ON key to return to the menu selection where you had previously left the instrument.
Sub-menu data entryIf you have selected a submenu (by pressing the
key shown in the prompt on the bottom of the display), but have not entered
in new data, then you must either:
6A
UX
AUXILIARY MODE
There are seven main menu options in the AUXILIARY mode:
1.
INITIALISE
2.
LIN SEP
3.
STN SEP
4.
AUTOGRAV SETUP
5.
OUTPUT
6.
EDIT
7.
LOCK AUX
INITIALISE Menu
INITIALISE MENU
There are twelve submenus in the INITIALISE mode. These menus, if selected,
require you to input specific data.
1.
To enter into the INITIALISE submenus, press the ENTER key and a submenu appears:
Example:
INITIALISE
ENTER?
Press:
eE
NTER
key
ERASE MEMORY
ENTER?
To enter into the submenu level, follow the prompt that is shown on the
second line of the display.
3.
INITIALISE SUBMENUS
The steps involved to enter a submenu are shown starting from the INITIALISE
menu. If you are already in the submenu level, then follow the steps starting
from step 2.
Parameters
2.
Instrument Parameters
1. ERASE MEMORY
At the start of each survey day ensure that there is adequate storage space
for the days data. The amount of memory available is displayed on the right
hand side of the screen (or by pressing the INFO key). If there is insufficient
capacity available, then erase the memory.
Erasing the memory does not erase initialization, setup or position parameters. However, all the recorded measurements in the memory will be lost.
Please ensure that you have transferred the data to your computer or printed
them out.
To erase the memory, do the following:
1.
2.
ERASE MEMORY
ENTER?
3.
ERASE MEMORY
START?
4.
5.
MEMORY FREE
TESTED 48K
100%
INITIALISE Menu
The 48K represents the size of memory installed in the Autograv. If the
value in the display does not agree with the installed value, then there
has been a partial memory failure.
2. JOB NO.
This submenu allows you to set a job number as shown in the following
steps:
1.
2.
JOB NO.
CHANGE?
3.
4.
5.
Parameters
3. OPERATOR
This submenu allows you to set an operator number as shown in the following steps:
1.
2.
OPERATOR
CHANGE?
Instrument Parameters
e
e
3.
4.
5.
4. GRID
This submenu allows you to set a grid number, as shown in the following
steps:
Note: You can also set this parameter by:
pressing the
INFO key on the keypad,
scrolling through the menu to the GRID prompt
and following the same steps as outlined below in steps 3 to 5.
1.
2.
GRID
CHANGE?
3.
4.
5.
5. LINE DIR
This submenu enables you to set the direction in which the lines are running
on the grid. You can choose from the following selections:
E/W
N/S
+/-
INITIALISE Menu
Example: If you select E/W, then a line is identified as north or south
of zero and a station (along a specific line) as east or west
of zero.
To select a line direction, do the following:
1.
2.
LINE DIR
CHANGE?
3.
4.
Press the
display.
5.
6. LINE
You can enter the line number that is to be surveyed first on the survey day.
Parameters
Note: The line and station data are recorded as positive and negative
numbers in the memory. The selection E/W or N/S is a filter
which only modifies the presentation of the data for display or
output. This is done by assigning either N or E for the positive
lines (or stations) as specified by the current setting. For example, a station record as +100 will be output as 100E, if the
line direction setting is not changed. However, if the setting is
changed to N/S before the data is output, then the same station will appear as 100N.
Instrument Parameters
1.
2.
LINE
CHANGE?
3.
4.
5.
7. STATION
At the start of a survey day you should enter in the first station number before taking a measurement.
1.
2.
STATION
CHANGE?
INITIALISE Menu
e
e
3.
4.
5.
1.
2.
3.
4.
Press the
When the measurement for a station has been recorded, the station number
is displayed along with the prompt Change?. In the lower line of the display, the number of the station where the measurements were just taken is
shown. Depending upon whether the Auto Station Increment was selected
or disabled, the upper line displays either the next station number or the
one just measured, respectively.
Parameters
Instrument Parameters
you must ensure that the station separation parameter (STN SEP)
reflects the direction of travel. During a survey, you might walk up one
line and down the next. To accommodate this change of direction, the
station separation must be changed also.
Example: When walking north (or east, or +), a station separation of
100N (or E, or +) will increment the station number by 100
with each recorded measurement. Walking down the next
line, you will be traveling south (or west, or ). For the automatic incrementation to work properly, the station separation must be set to 100S (or W, or ) so that the station
numbers will now decrement.
5 0
Stations:
Line
Direction
N/S
E/W
+/-
Press
Press
north of zero
increments
north of zero
decrements
south of zero
decrements
south of zero
increments
east of zero
increments
east of zero
decrements
west of zero
decrements
west of zero
increments
+ zero increments
+ zero decrements
- zero decrements
- zero increments
INITIALISE Menu
If you select +/ for the line direction, pressing
station number, possibly to a negative value.
Notes:
0 decreases the
9. RECAUTOSTART
This feature must be disabled to operate the Autograv.
To disable the RECAUTOSTART feature, do the following:
1.
2.
REC->AUTOSTART
CHG?
>nnnnn<
3.
4.
Press the
10. SERIAL #
The Autograv serial number should be entered into the memory. The serial
number is located on the bottom of the case. Please enter the six right-most
digits, as the Autograv will only accept six characters.
To enter the serial number, do the following:
1.
2.
Parameters
Instrument Parameters
SERIAL#
CHANGE?
3.
4.
5.
12. HEATER
If the Heater function is selected, then the display heater automatically turns
on when the temperature drops below -20C.
1.
2.
HEATER
CHG?
>nnnnn<
3.
4.
Press the
INITIALISE Menu
2.
3.
TIME:
CHANGE?
XX:XX:XX
XX:XX:XX
TIME:
ENTER?
HH:MM:SS
XX:XX:XX
4.
Key in the hours and press the ENTER key to go to the next parameter in
the time field.
5.
Press the ENTER key after you have typed in the seconds.
TIME:
CHANGE?
e
e
HH:MM:SS
YY:MM:DD
6.
Key in the year and press the ENTER key to go to the next parameter in
the date field.
7.
Press the ENTER key after you have typed in the day. When you enter in
the day, the clock starts to run.
Parameters
e
e
1.
Instrument Parameters
LINE SEP
There are no submenus in the LINE SEPARATION mode. This mode requires
the entry of the regular distance between survey lines in metres or feet.
During a survey, when you change to the next line, the instrument advances
its records to the next line number when you press the
LINE and
(or
) keys.
1.
Display the LIN SEP prompt and press the CHANGE key. The following prompt appears:
LIN SEP
ENTER?
2.
3.
STN SEP
STN SEP
There are no submenus in the STATION SEPARATION mode. This mode requires the entry of the distance between stations in metres or feet. During
the survey, when you change to the next station, the instrument advances its
records to the next station number.
If the direction of travel has changed and Automatic Station Incrementation
is selected, then the sign (or direction) of the separation parameter has to
be changed.
To set the station separation value do the following:
1.
Display the STN SEP prompt and press the CHANGE key. The following prompt appears:
STN SEP
ENTER?
3.
Parameters
2.
Instrument Parameters
AUTOGRAV SETUP
There are 14 submenus in the AUTOGRAV SETUP mode. Four of the submenus appear only if specific parameters are selected. A second level of
submenus is also available for only some of the first level submenus.
1.
To enter into the first level submenu of the AUTOGRAV SETUP, press the
ENTER key and a submenu appears:
Example:
AUTOGRAV SETUP
ENTER?
Press:
e (E
NTER)
key
MODE:
CHANGE?
NNNN
2.
To enter into the second level of submenus, follow the prompt that is
shown on the second line of the display.
3.
AUTOGRAV SETUP
1. MODE
You can select from three modes of operation:
FLD GRAV
CYCLING
NONE
1.
Display the AUTOGRAV SETUP prompt and press the ENTER key.
2.
MODE:
CHANGE?
3.
4.
Press the
display.
5.
NNNN
1.
Display the AUTOGRAV SETUP prompt and press the ENTER key.
2.
Parameters
Instrument Parameters
READ TIME
CHANGE?
3.
4.
5.
XX
If you have selected the FLD GRAV or NONE mode, then go to step 10 as the
CYCLE TIME prompt will not appear in the display.
If you have selected the CYCLING mode, then:
6.
Press the
e
Notes:
7.
8.
XXXX
If any of the DATA, INFO or SCROLL keys are pressed, then the
display will appear to be locked-up. You must then press and hold
the ON key for more than 5 seconds to reset the instrument. If any
other key is pressed while the meter is in the OFF stage of the
CYCLE mode, then you must press the STOP key to stop cycling.
AUTOGRAV SETUP
e
6
9.
a) CHART SEL:
This submenu appears only if you select the CYCLING mode. You can
choose the full scale range of the output of the digital to analogue
converter from the following selections:
0.1 mGal
1.0 mGal
10.0 mGal
100.0 mGal
Note: See page 63 for details about using the chart recorder.
To set the chart select do the following:
2.
CHART SEL:
CHANGE?
e
e
5 or 0 key
XX
3.
4.
Press the
or
key until the desired chart selection
appears on the display.
5.
5 0
Parameters
1.
Instrument Parameters
1.
Display the AUTOGRAV SETUP prompt and press the ENTER key.
2.
3.
4.
ERR. LIMIT
CHANGE?
5.
6.
7.
XXXX
AUTOGRAV SETUP
4. CAL. AFTER
You can set the internal calibration frequency with this value. Please refer
to page 42 for more information about the CAL. AFTER feature.
To set this function do the following:
1.
Display the AUTOGRAV SETUP prompt and press the ENTER key.
2.
CAL. AFTER
CHANGE?
3.
4.
Key in the new calibration after value (seconds). (Refer to page 42 for
recommended value).
5.
1.
Display the AUTOGRAV SETUP prompt and press the ENTER key.
2.
Parameters
Instrument Parameters
3.
4.
Press the
Note: If the Continuous Tilt Correction feature is selected in a previous reading, a small square appears on the top line of the display next to GRAV.
6. AUTO REJECTION
You can set the Autograv to automatically reject noisy samples (page 44)
as follows:
1.
Display the AUTOGRAV SETUP prompt and press the ENTER key.
2.
AUTO REJECTION
CHG?
>NNNN<
3.
4.
Press the
7. TIDE CORRECTION
You can set the tide correction feature using this function as follows:
1.
Display the AUTOGRAV SETUP prompt and press the ENTER key.
2.
AUTOGRAV SETUP
TIDE CORRECTION
CHG?
>NNNN<
3.
4.
Press the
If you select the tide correction feature the following three submenus
appear:
DEG. LAT
DEG. LONG
GMT DIFF
a) DEG. LAT
This submenu appears only if you select the tide correction feature.
1.
2.
DEG. LAT
CHANGE?
e
e
3.
4.
5.
5 or 0 key
XX.X
Parameters
Instrument Parameters
b) DEG. LONG
This submenu appears only if you select the tide correction feature.
To set the degrees latitude do the following:
1.
2.
DEG. LONG
CHANGE?
e
e
c)
3.
4.
5.
5 or 0 key
XX.X
GMT. DIFF
This submenu only appears if you select the tide correction feature.
To set the Greenwich mean time (GMT) difference do the following:
1.
2.
GMT. DIFF
CHANGE?
3.
5 or 0 key
X
AUTOGRAV SETUP
4.
Key in the value. (GMT - clock time) in hours. This will be negative
east of Greenwich and positive west of Greenwich.
5.
8. AUTO-RECORD
You can set the instrument so that it automatically records the results of a
reading without displaying it on the screen.
To set the Auto-record feature do the following:
1.
Display the AUTOGRAV SETUP prompt and press the ENTER key.
2.
AUTO-RECORD
CHG?
>NNNN<
3.
4.
Press the
9. GREF.
The GREF value is the gravity offset added to the gravity reading.
To set the GREF value do the following:
1.
Display the AUTOGRAV SETUP prompt and press the ENTER key.
2.
Parameters
Instrument Parameters
GREF:
CHANGE?
3.
4.
5.
>
a) Drift
To set the drift correction do the following:
1.
5 or 0key
DRIFT CORRECT.
ENTER?
2.
3.
DRIFT:
CHANGE?
>
XXX.
AUTOGRAV SETUP
e
e
4.
5.
6.
1.
2.
5 or 0 key
DRIFT CORRECT.
ENTER?
3.
4.
Press the
5.
Press the ENTER key and the prompt shown at the top of the
display shows you the current start time.
6.
The prompt at the top of the display changes to the current real-time
clock setting.
Parameters
Instrument Parameters
c)
2.
5 or 0 key
DRIFT CORRECT.
ENTER?
3.
4.
Press the
eee
5.
Repeatedly press the ENTER key until the prompt shown at the top
of the display reaches the desired start time.
6.
AUTOGRAV SETUP
1.
Display the AUTOGRAV SETUP prompt and press the ENTER key.
2.
INIT. CONST.:
CHANGE?
4.
GCAL.1:
CHANGE?
c.
XXX
5.
d. Press the
key to go to the next initial constant.
Repeat step 4 until you have entered all five initial constant values.
Parameters
3.
Instrument Parameters
6
OUTPUT
6.
The measured data is stored in the internal memory and dumped in ascending order according to the:
grid
line
station
time values (in this order of priority).
There is no way to change this order, i.e. the measurement recorded at a station with the lower station number will be dumped before the measurement
at the higher station number, regardless of the time it was taken.
There are seven submenus in the OUTPUT mode.
1.
To enter into the OUTPUT mode, press the ENTER key and a submenu
appears:
Example:
OUTPUT
ENTER?
Press:
eE
NTER
key
AUTOGRAV
START DUMP?
2.
To enter into the second level of submenus, follow the prompt that is
shown on the bottom line of the display.
3.
OUTPUT
Notes:
OUTPUT SUBMENUS
The steps involved to enter a submenu is shown from the OUTPUT mode. If
you are already in the submenu level follow the steps from step 2.
1.
2.
3.
q
w
c.
Parameters
AUTOGRAV
START DUMP?
Instrument Parameters
3. PLOT
This submenu enables you to set the characteristics of the plot by setting six
of the seven second level submenu parameters. The plotting is simulated by
a character position on the print line giving you a simplistic graphic presentation. One line is used per data record.
Two different sets of data can be traced at one time.
The PLOT submenu has a further second level submenu for setting the
following parameters:
TRACE1
BIAS1
SCALE1
TRACE2
BIAS2
SCALE2
START PLOT
1.
2.
PLOT
ENTER?
a) TRACE1
3.
OUTPUT
TRACE1:
>
SELECT/ENTER
NNNN
You must select one of the following parameters as your first trace:
INFO A
GRAV
INFO B
TILT X
INFO C
TILT Y
INFO D
TEMP
INFO E
SD.G
INFO F
# REJ
INFO G
CAL AFTER
INFO H
a. Press the ENTER key.
c.
b. Press the
5 0
or
key until the desired characteristic
appears on the display.
Press the ENTER key again.
b) BIAS1
4.
XXXX
Parameters
Instrument Parameters
To enter a BIAS value:
c.
c)
SCALE1
5.
Press the
6.
If you press the CHANGE key, then the following prompt appears:
Exponential Value
XXXX
<E X
OUTPUT
Parameters
Autograv CG3 Manual 3-35
Instrument Parameters
d) TRACE2
7.
Press the
8.
e) BIAS2
9.
NNNN
Press the
XXXX
10. You set the second Bias value as outlined in BIAS1 on page 3-33.
f)
SCALE2
12. You set the second scale as outlined in SCALE1 on page 3-34 in step 6.
OUTPUT
g) START PLOT
13. Press the
q
w
c.
4. REMOTE
5. BAUD RATE
You must set the baud rate that the instrument sends data. The fastest operating baud rate for the Autograv is 2400 baud. There are five different baud
rates in the menu:
2400
1200
600
300
110
1.
2.
Parameters
This function is for SCINTREX engineering use only. Please ignore it.
Instrument Parameters
BAUD RATE:
CHANGE?
XXX
3.
4.
5.
6.
If the 2400 baud rate is selected and the printing does not start, then you
must press and hold the
ON key to clear the apparent hang-up.
6. DATA BITS
A choice of 7 or 8 data bits is available.
To set the data bits do the following:
1.
2.
DATA BITS:
CHANGE?
3.
4.
Press the
5.
OUTPUT
7. CR DELAY
You can set the carriage return (CR) delay according to the printer speed.
It is a value between 0 and 999.
Notes:
1.
2.
CR DELAY:
CHANGE?
3.
4.
5.
Parameters
5 or 0 key
Instrument Parameters
EDIT
You can do the following three types of changes to the data in memory via
the EDIT mode:
Erase a measurement
Change a station, line or grid number, or ancillary information
Repeat a measurement for a particular station or line
Warning: Once you enter the EDIT mode, filed data can be permanently
changed or accidently lost. Scintrex Limited does not
recommend that you use the EDIT mode. It is preferable to edit
the data that has been dumped to the computer. For more
information contact Scintrex Limited.
LOCK AUX.
LOCK AUX.
The LOCK AUX mode enables you to lock the AUXILIARY mode in order to
prevent unintentional change of system parameters during surveys.
When the battery is connected or if the CG-3 is reset, the LOCK AUX function
is automatically disabled. All of the system menu options will appear in the
display.
5 or 0 key
LOCK AUX.
CHG? >DISABLED<
2.
3.
Press the AUX. key twice. All of the AUXILIARY menu options
disappear from the display except Station Separation (STN
SEP).
4.
Parameters
e
66
Instrument Parameters
To reset the instrument, press and hold the
while turning on the Autograv.
w (O /O
N
Autograv CG3
Section
FIELD OPERATION
Several features of the Autograv, such as the integrated housing with the
built-in shock mounts and the absence of an internal clamping system,
make it easy to safely transport the instrument and to obtain accurate
readings.
The following pages describe typical field operating procedures for the
Autograv.
3. Push the START key once to get into the ADJUST mode.
Field Operation
Field Operation
The Firmware section and Appendix A: Theory of Operation give full descriptions of the Autograv software function. This section describes how to select
setup parameters and software settings for field operation.
CAL. AFTER
As described in Appendix A: Theory of Operation, the CAL.AFTER parameter adjusts the ratio of internal calibration samples to gravity samples. Reducing the ratio (reducing the frequency of internal calibration) increases
the efficiency of the reading process as more gravity samples are being accumulated in a given time. However, the internal electronic noise increases
as the number of ADC calibrations is reduced.
1
NOTE:
The minimum reading time for readings terminated on the ERR. LIMIT criterion is 15 seconds.
Field Operation
Field Operation
RECORDING DATA
The way data is recorded depends on whether you select or disable the AUTO-RECORD option in the AUTOGRAV SETUP.
If you disable the AUTO-RECORD feature, you must press the RECORD
key when a reading is terminated.
If you select the AUTO STN INC option in the INITIALISE menu, the
station number is automatically changed by an amount equal to the
STN SEP entry in the AUXILIARY mode menu.
If you disable the AUTO STN INC option, you can change the
station number by a keyboard entry or by using the SCROLL keys to
change the station setting by an amount equal to the STN SEP entry.
AUTO REJECTION
This feature eliminates noise spikes from a reading and should be selected
during field surveys. Samples which are more than 4 standard deviations
away from the average are rejected. No rejections are made from the first five
samples which are used to build up an average and standard deviation.
While the measurement progresses, the standard deviation is continuously
recalculated and the most recent value is used to set the rejection thresholds of 4 standard deviations around the summing gravity average. The
standard deviation (SD) of the 1 second gravity signal sample is displayed
during calibration measurements, i.e. every 12 seconds, if CAL AFTER is
set to 12.
Therefore, if you can identify sources of noise spikes, start the reading in a
quiet time so that a realistic average and standard deviation can be built up
to serve as a basis for rejections. For example, if you want to take a reading
beside a road, wait until there are no cars going past before you start the
reading.
Recording Data
The auto-rejection feature is also quite useful in the situation where the instrument is cycling and connected through the battery charger to the AC
power line. The voltage supplied could be noisy and cause the occasional
spikes in readings.
NOTEBOOK FEATURE
With each measurement that you take, you can enter in up to eight sections
of coded information (Info A Info H). Within each section, up to 5
digits may be entered. To enter in information, do the following:
1. Before a measurement has been recorded, press the INFO key and scroll
Field Operation
Autograv CG3 Manual 4-5
Field Operation
TRANSPORTING
AND
HANDLING
The Autograv case is waterproof and the instrument has been designed to
be transported between stations without the dust cover. The dust cover only
needs to be installed when the instrument is being shipped or when there
is a danger of the front panel or control console being damaged.
To get the best results from your Autograv and to ensure a long working life,
treat your instrument carefully:
Avoid shocks and jolts caused by hitting the instrument against rocks
and trees etc., when taking the instrument out of vehicles and placing
it on the tripod.
OPERATION HINTS
The following are some hints for better instrument results:
Operation Hints
Follow the same setup and reading routine at each station. For
example, if you are transporting the Autograv by vehicle, take each
station reading at approximately the same time interval after the
vehicle has stopped.
TRANSPORTATION
The CG-3 is an extremely sensitive and precise scientific instrument. It
is capable of measuring gravity changes which would result in 10-10
metres change in the proof mass vertical position. Note that the hydrogen atom has a diameter of the same size. However, it is inherently
robust and well capable of providing accurate relative gravity values
under relatively rough (but normal) field handling. In a controlled
shock test, one production meter was dropped from various heights to
a hard floor. The drops from 40mm, resulting in a 20g peak deceleration at the sensor location, resulted in level shifts of less than 0.01
mGal. The shock isolators and the stops limiting the proof mass movement, reduce the effect of shocks on the sensor. No CG-3/3M sensors
have been found to be broken since production started in the second
half of 1989. Nevertheless, it too has limits, so that careful handling during transportation generally yields better measurement results.
Field Operation
Field Operation
the first station. By observing the sequential readings (e.g. 60 or 120
seconds averaging) you can tell if the unit is stable by judging their
convergence.
MOTION NOISE
Motion of the base of support of the meter may induce noise and consequent reading errors. Such motion may be caused by:
Poor ground conditionsSoft ground, tree root movement and settling
of the tripod. These are improved by the Tilt Correction (continuous or not)
which is described on page 4-3.
Microseismic noise This is due to motion of the continental crust as
an effective wave guide for seismic surface waves of (usually) a 5 to 7 second period. These effects may be caused by long-period ocean waves striking coast lines. They are most severe in islands in the seas (e.g. Ireland), but
may at times be troublesome even in the centre of large continents.
The effect of microseismic noise may best be reduced by measuring for larger periods and thus achieving the required statistical accuracy of the measurement.
Earthquake shock wavesThe shock waves from an earthquake may
affect measurements so much that they cannot be carried on while these
shock waves last.
Industrial traffic noise Vibration due to industrial activity and/or
the nearby movement of heavy vehicles generally introduces higher frequency seismic noise (e.g. 4 to100Hz). Gravity stations which are near such
sources of noise should be avoided. Where they cannot be moved to a quieter location, measure longer, e.g. 120 sec instead of 60 sec.
The standard deviation of 1 second measurement samples is a a good indication of the seismic noise, especially longer period microseismic noise.
During quiet periods, SD is in the range of 0.010 to 0.030 mGal, whereas, during storms on open seas, it could increase to 0.150 mGal (measured in Toronto) or even higher near the sea coast. It could go to 0.5 mGal and higher
during remote but strong earthquakes. The shorter period industrial noise
Operation Hints
is reduced by signal averaging over 1 second so that the 1 second samples
are less responsive to this noise. Noise spikes are removed by auto rejection
of 1 second samples.
WIND
Wind striking a gravity meter can cause both vibration noise and reduction
in gravity by tilting the meter out of verticality. Tilting the meter so that it is
not vertical tends to reduce the apparent gravity value, despite the statistical
averaging of the gravity measurements and the automatic tilt correction.
A windbreak should be devised to protect the meter from the effect of wind.
A simple windbreak can be made from a sheet of canvas about 1.5m square,
fastened on each of two opposing sides to two aluminum tubes or wooden
rods (each, about 2m (6.5ft) long). The operator (or a helper) then holds
the rods outstretched and vertical on the upwind side of the gravity meter
to shield it from the wind. Alternatively, a large (golf) umbrella could also
be used.
have low vibration noise, e.g. away from industrial, vehicular or natural
vibration sources (rivers, beaches, volcanoes etc.) and
have a hard, level surface where the precise tripod location may be
repeated, and the level of the gravity meter maintained.
Field Operation
Field Operation
An external battery belt and extender cable (optional) can reduce the effect of a cold battery as the external battery is kept warm by the heat of the
operator (see page 5-2).
Tip:
Dumping Data
1. After the instrument has been powered up for 12 hours, check the Tem-
base station(s) and correct for the drift in the data reduction. Do not
attempt to adjust the DRIFT constant in the CG-3 software yet. For the
first day of surveying, it is advisable to re-occupy base stations approximately once every two hours. On the second day, once every 4 hours
is probably sufficient.
3. At the end of the first and second days work, check the Temperature
DUMPING DATA
Once the Autograv has been physically linked to a computer by means of
adapters and serial cables, it is necessary to make them communicate. The
interface software may be a commercially available communications
package, or may be custom software dedicated to the users hardware
configuration.
Field Operation
Field Operation
Autograv CG3
Section
LOSS
OF
POWER
The instrument must be powered at all times (even during storage) by either
the internal 12V battery or the battery charger. If a loss of power to the instrument occurs, then several things happen:
After a few seconds of power loss, the clock setting is erased and must
be reset to the current time after power is resumed.
The thermostat loses power and the instrument begins to cool down.
The consequences of the power loss depends upon the length of time
that the instrument is powered down.
Check the temperature output in the ADJUST mode to see when the
thermostat is stable. The stabilization time increases as the powereddown time increases.
Battery
a few minutes
30 minutes (approx.)
12 hours
24 hours (approx.)
48 hours or more
48 hours (approx.)
If the instrument is powered down for more than 48 hours, the contents
of the memory, including the Autograv initialization constants could be
lost. If this occurs, you will have to go through the Initialization procedures outlined in the Getting Started section.
BATTERY
The standard internal rechargeable battery for the Autograv is the
Panasonic12V/7.2Ah (previously, the Sonnenschein 12V/6Ah battery was
used). When this battery is fully charged, it provides enough power to operate the instrument throughout a normal survey day. However, battery lifetime increases as the ambient temperature increases. At 25C, the newly
installed battery lasts for 15 to 16 hours but at -10C, the battery lasts for only
seven hours.
TEMPERATURE EFFECTS
A typical plot of CG-3 service hours against the ambient temperature for the
Panasonic battery is shown in Figure 12 on page 5-3. At low temperatures,
the service hours decrease rapidly for two reasons:
To circumvent the first effect and extend the battery life under cold weather
conditions, an external battery belt (optional) can be provided. This battery
belt is strapped to the operator's waist underneath the clothing and thereby
kept warm. An extender cable to the charger plug on the Autograv provides
5-2 Autograv CG3 Manual
Battery
power to the instrument. Once the external battery supply has been connected, the internal battery, if discharged, should be removed, as it should
not be kept in place for very long thereafter. When the internal battery has
been replaced, the external battery may be then disconnected without any
interruptions of the temperature stabilization of the sensor.
Figure 12
VOLTAGE
Low battery warningIf the battery voltage drops below 10.5 volts, the
Autograv emits an audible low battery level alarm. The battery has approximately 1 hour left of operating time after the alarm sounds. The battery
should then either be recharged by connecting the instrument to the charger or replaced with a charged battery.
Battery
LIFE EXPECTANCY
The life expectancy of the battery is approximately 200 cycles (each cycle
being a complete discharge and recharge). In standby operation, the life expectancy is 4 to 5 years. Two batteries are supplied with each Autograv.
If the CG-3 battery has been in use in the field for 150 to 200 days, then its
capacity to hold a charge may be significantly less than it was new. At this
point, it is more economical to replace the battery with a new one, rather
than suffer with the operational inconvenience of a reduced battery life and
frequent battery changes in the field. Figure 13 on page 5-4 illustrates the typical deterioration of the battery capacity with the number of discharge/
charge cycles.
Charger
REPLACING
You should replace both of your batteries with new ones, approximately
once a year if your CG-3 has been in relatively steady use. You should replace
an individual battery even earlier, if you should find that its capacity has
been reduced, even after it has been fully charged.
When replacing a CG-3 battery, it is recommended that you do so only with
one of a type which has been tested and approved by Scintrex. To date, the
only recommended batteries are:
The reason why a general approval cannot be given for all 12V lead-acid batteries is that each brand may have different charging requirements, resulting
in incompatibility with the CG-3 charger.
Hints:
1.
2.
3.
Replace a battery after about 200 days of field use, as it will no longer
provide a proper lifetime in the field.
CHARGER
The charger serves three functions:
a.
b.
c.
Battery
CHARGING TIME
Charging time for batteries is approximately eight hours. For best charge acceptance by the batteries, the ambient temperature should be in the 20 to
30C range, wherever possible.
115V
230V
When you are selecting a line voltage you must set the charger voltage
switch to the correct line voltage.
2.
Using a small flathead screwdriver, flip the recessed voltage switch that
is on the front of the charger to the appropriate setting.
Charger
3.
Fuse
115V
0.75A
230V
0.375A
When you plug the charger into an AC outlet, the green light turns on to indicate that the charger is working.
Notes:
To operate the Autograv with the 230V setting, you need the optional
230V adaptor plug.
If you use the 0.75A fuse with the 230V setting, then the battery and the
charger will not be protected against potentially dangerous fluctuations in the line voltage, or the electronic component failure.
If you use the 0.375A fuse with the 115V setting, then you will blow the
fuse when you plug the charger into an AC outlet.
The battery charger gets hot when in use. This is normal and the
charger has an automatic thermal limiting circuit to reduce power dissipation before overheating occurs. To reduce heating, ensure that the
charger is uncovered in an open place and that air flow over the case
and fins is not restricted.
If the instrument has a discharged battery in it when the charger is connected, then wait for 1 to 2 hours before plugging in another discharged battery into a side socket.
Battery
HOW
TO
CHANGE
BATTERY
1.
2.
Without unplugging the discharged battery, remove the battery from the
battery compartment by pushing on the left side of the battery and then
pulling the black tab towards you so the battery can slide out (see
Figure 15 on page 5-9).
Insert the connector of the new battery into the spare socket on the left
hand side of the battery compartment wall (see Figure 15).
4.
5.
Before you reinstall the new battery, place your hand inside and flatten
the black tab against the bottom, side and top of the compartment. If
the black tab is not pressed against the walls of the compartment, then
the battery will not fit and the foam padding will be damaged.
6.
Battery
Autograv CG3 Manual 5-9
Autograv CG3
Section
Temperature compensation
Drift correction
Tilt correction
Calibration constant
TEMPERATURE COMPENSATION
Important: The temperature compensation must be checked and
adjusted (if required) before the drift correction.
The reason for apparent long term drift of the monitored chamber temperature is the slow aging of thermistors. The thermistors are used for the
sensor chamber thermal stabilization and the chamber temperature
monitoring and compensation. The measurement range of the temperature monitoring circuits is 3 mK. It is advisable to readjust the reading
offsets well before the ends of the range are reached.
Check the temperature compensation if the instrument has been turned
off for more than two hours and also periodically during field operations.
Maintenance
MAINTENANCE
Maintenance
With the instrument at room temperature, the temperature reading must be
within the range of: -1.0 mK < TEMP < +1.0 mK
Please refer to the Getting Started section on how to check and adjust the
temperature compensation (starting on page 2-35). Also, note that the
adjustment does not change the sensor chamber temperature, but only the
temperature reading offset circuits. Consequently, the gravity readings will
differ by the value TEMP x TEMPCO, where TEMP is the temperature
reading change and the TEMPCO is the software correction constant (see
page A-15). The previous level would be re-established by using the GREF
function, see page 3-25.
DRIFT CORRECTION
1.
During the first month of operation for a new instrument, check the drift
correction once a week.
2.
After this four week period, check the drift correction once every
month or longer depending on the individual characteristics of the
instrument.
Important: If an instrument has been turned off for several weeks or
more (which should never be done without a valid reason), then the setup procedure outlined in Section 2
must be followed as if the instrument was new.
3.
Reset the drift correction start time (refer to page 2-41) at the same
time as the new drift constant is entered. The drift correction start time
must be reset before the difference between it and the current time
exceeds 4 months. (Refer to equation 9.8 on page A-13 in Appendix A)
The drift correction should not be adjusted on the basis of repeat measurements at base stations during a field survey, but rather with drift
measurements obtained from a stationary instrument. The drift correc-
Drift Correction
Ensure that the tide correction is selected and properly adjusted while
measuring the drift correction.
Maintenance
tion feature corrects for relatively constant long term instrument drift
due to stress relaxation in the elastic system. There may be a small
amount of additional drift induced by transportation and handling.
The drift correction is not intended to eliminate drift due to transportation as it is generally not constant and depends on the conditions of
transportation. This residual drift can be determined by repeated measurements at selected base stations.
Maintenance
Measuring XE
1.
Setting
MODE
FIELD
READ TIME
120
TILT CORRECTION
SELECTED
TIDE CORRECTION
SELECTED
AUTO REJECTION
SELECTED
DISABLED
3.
Press the START key. The TILTS value appears on the display.
4.
5.
Wait for 5 minutes and then note the exact value (X1) as shown on the
display.
6.
7.
8.
Wait for 5 minutes and then note the exact X value (X2) as shown on
the display.
9.
10.
where X1, X2 and XE are in units of arc sec and R1, R2 are in mGal.
Example:
X1 = +156
X2 = 149
R1 = 4236.58
R2 = 4236.54
4236.54 4236.58
X E = ------------------------------------------------- 43386
156 ( 149 )
0.04 43386X E = ------------------------------------305
Maintenance
2.
Maintenance
Fine Adjustment of XE
If XE 0, the Autograv must be adjusted so that the non-level readings are
correctly compensated.
1.
2.
3.
Press the START key to put the instrument in the ADJUST mode.
4.
5.
Example:
Maintenance
Maintenance
Measuring YE
The following procedure is basically the same as that for measuring XE.
1.
2.
3.
Wait for 5 minutes and then note the exact Y value (Y1) as shown on
the display.
4.
5.
6.
Wait for 5 minutes and then note the exact Y value (Y2) as shown on
the display.
7.
8.
where Y1, Y2 and YE are in units of arc sec and R1, R2 are in mGal.
Example:
Y1 = +156
Y2 = 149
R1 = 4236.58
R2 = 4236.54
4236.54 4236.58
Y E = ------------------------------------------------- 43386
156 ( 149 )
0.04 43386YE =
------------------------------------305
Fine Adjustment of YE
If YE 0, the Autograv must be adjusted so that the non-level readings are
correctly compensated.
Leave the instrument in the Y = Y2, X = 0 orientation.
2.
3.
Press the START key to put the instrument in the adjust mode.
4.
Y2' = Y2 + YE
Maintenance
1.
Maintenance
Coarse Adjustment of XE
1.
2.
3.
4.
5.
6.
Carefully rotate the X tilt sensor coarse zero adjustment until X = XE.
Note: Potentiometer sensitivity is 2000 arc seconds/turn and adjusting the pot. counter-clockwise increases the reading. Carefully
rotate the screw using very small increments while viewing the
liquid crystal display.
7.
8.
Coarse Adjustment of YE
Remove the Coarse Zero Adjust cover.
2.
3.
4.
5.
6.
Carefully rotate the Y tilt sensor coarse zero adjustment until Y=YE.
Note: Potentiometer sensitivity is 2000 arc seconds/turn and adjusting the pot. counter-clockwise increases the reading. Carefully
rotate the screw using very small increments while viewing the
liquid crystal display.
7.
8.
Maintenance
1.
Maintenance
Setting
MODE
FIELD
READ TIME
120
TILT CORRECTION
SELECTED
TIDE CORRECTION
SELECTED
AUTO REJECTION
SELECTED
DISABLED
2.
Place the Autograv on the tripod in a quiet location with a solid floor
(concrete is preferable). A wooden floor or carpet is not suitable.
3.
Press the START key. The TILTS value appears on the display.
4.
5.
6.
7.
Wait for 5 minutes and then note the exact value (X1) as shown on the
display.
8.
K =
R0 R1
4
1 + 8.7 10 -------------------X1 2
Example:
2.
3.
X1 = +155
TILTXS = 195.3
R0 = 4236.52
R1 = 4236.56
therefore,
TILTXS =
0.04
4
- 195.3
1 + 8.7 10 ------------- 155 2
TILTXS = 180.6
Maintenance
where
Maintenance
Setting
MODE
FIELD
READ TIME
120
TILT CORRECTION
SELECTED
TIDE CORRECTION
SELECTED
AUTO REJECTION
SELECTED
DISABLED
2.
Place the Autograv on the tripod in a quiet location with a solid floor
(concrete is preferable). A wooden floor or carpet is not suitable.
3.
Press the START key. The TILTS value appears on the display.
4.
5.
6.
7.
Wait for 5 minutes and then note the exact value (Y1) as shown on the
display.
8.
K =
R0 R1
4
1 + 8.7 10 -------------------Y1 2
Example:
2.
3.
Y1 = +155
TILTYS = 195.3
R0 = 4236.52
R1 = 4236.56
therefore,
TILTXYS =
0.04
4
- 195.3
1 + 8.7 10 ------------- 155 2
TILTYS = 180.6
Maintenance
where
Maintenance
Measuring
1.
Place the Autograv on the tripod in a quiet location with a solid floor
(concrete is preferable). A wooden floor or carpet is not suitable.
2.
Rotate the tripod footscrews eight turns up from their bottom positions.
3.
Press the START key to set the Autograv into the ADJUST mode and level
the instrument until X = 0 10, Y = 0 1.
4.
Wait 5 minutes and observe the Y-tilt. Record the value as Y0.
5.
Rotate the front footscrews clockwise five turns. Observe the Y-tilt and
record the value as Y1.
6.
7.
If this specification is not met, then proceed with the steps as outlined in the
following Adjustment section.
Adjusting
Place the Autograv on the tripod in a quiet location with a solid floor
(concrete is preferable). A wooden floor or carpet is not suitable.
2.
On a piece of protective foam, lay the Autograv down on its front and
view the bottom of the instrument. See figure below:
Loosen the four screws on the Vee groove plate (slot side) until the
plate can just slide freely.
4.
Raise the tripod feet eight (8) turns from their bottom positions.
Maintenance
1.
Maintenance
5.
6.
Press the START key to place the instrument into the ADJUST mode.
7.
8.
Wait 5 minutes and observe the Y-tilt. Record the value as Y0.
9.
Rotate the front footscrew clockwise five turns. Observe the Y-tilt output.
10. Hold the tripod in place as it must not move during this step. Rotate the
instrument on the tripod until Y = Y0 5 (if Y < 0, then rotate instru-
ment clockwise). The Vee groove block should slide relative to the case.
If possible, have a second person hold the tripod.
11. Tighten the four screws on the Vee groove plate.
12. Repeat steps 8 and 9. The difference in Y (i.e. Y0 Y) should be less
than 5 seconds. If it is not within this range, then repeat the adjustment
procedure.
RECALIBRATING
THE
GRAVIMETER
As explained on page A-10, the calibration factors GCAL1 and GCAL2 are
applied to the gravity signal, according to the Equation (A.4) on page A-10.
The calibration factor GCAL2, which accounts for a small quadratic nonlinearity, is evaluated in the Scintrex gravity laboratory. Since 1991, it has been
reduced to zero by an electronic adjustment. The GCAL2 need not be reevaluated.
The main calibration constant, GCALI, is determined on the Orangeville
Calibration Line, 70 km north of Toronto. This line, which includes two absolute stations, was established and is maintained by the Geological Survey
of Canada. The section used for the CG-3 calibration and testing consists of
5 stations spread over a distance of 70 km covering a gravity interval of 106
mGal (recently modified to 119 mGal). Instruments are transported by car,
strapped onto the back seat with no special transport case or fixtures.
Readings are taken at each station on the way up the line and repeated on
the way down. Readings of DUR=60 seconds are taken. Earth tide corrections are made automatically with the instruments built-in program.
where gij are the reference gravity differences and k is a constant (the scale
correction factor) which is varied until the standard deviation of the measurements errors is minimized. The new scale factor GCAL1 is then calculated from the old value GCAL1 used during the calibration test as
GCAL1 = k * GCAL1
The accuracy of the GCAL1 determination is about 0.01% (0.01 mGal measurement error in 106 mGal gravity test range).
The GCAL1 stability depends on the dimensional stability of the capacitive
displacement transducer, and the stability of the internal D.C. reference voltage. After the initial period of a few months, during which the GCAL1 may
change, due to the stress relaxation effects in the newly fused quartz sensor,
by up to 0.1%, the drift rate of the scale factor GCALl is commonly 1 to 2 ppm
(parts per million) per day. After a few years, users requiring greatest precision, in particular for regional surveys, may wish to recalibrate their
gravimeters. This could be done in a similar way as it was done initially at
Scintrex.
In order to derive the new GCALl at least two locally available stations with
the reliably determined gravity difference of at least 100 mGal are necessary.
Additional intermediate well determined gravity stations will serve to increase the precision of this determination. By looping back to the starting
station the residual drift should be determined, and used for correction of
readings at other stations. Beware that some older calibration ranges were
established by using LaCoste-Romberg gravimeters before the existence of
cyclic measurement screw errors was known and the corrections for it applied, and that therefore they may not be very accurate.
Maintenance
Maintenance
TRIPOD MAINTENANCE
The tripod footscrew adjustment threads are protected by a stainless steel
sleeve at the tripod base and a plastic wiper in the bottom of each footscrew.
The stainless steel sleeve should be periodically maintained by winding the
tripod footscrews to the top of their travel, and then cleaning and lightly oiling the sleeve.
STORING
THE
AUTOGRAV
There are several steps that you must follow in order to obtain the best results
from your instrument:
Store the Autograv with the powered up battery charger in a dry, secure
place.
Autograv CG3
Section
TROUBLESHOOTING
Problems
Possible Causes
Possible Remedies
Battery is depleted.
Display is stuck.
Reading appears to
be out of range or
reading is close in
value to GCAL1 and
ERR/SD is low.
1.Gently tap the upper panel underneath the Autograv name with your
finger several times.
2.Take a new reading so that the
instrument does not average the
incorrect reading.
3.If the reading is still incorrect, repeat
the steps increasing the number of
taps until the sensor unlatches.
Autograv CG3 Manual 7-1
Troubleshooting
The following is a brief outline of problems, possible causes and possible remedies to aid you
in the event of system failures. Follow the natural progression of steps that is shown below for
troubleshooting the Autograv.
Troubleshooting
Problems
Possible Causes
Possible Remedies
Display changes
slowly
Ambient temperature is
too cold for the display to
operate properly
Miniature batteries on
memory board need
replacement
Auxiliary Mode
locked
Instrument needs to be
reset
1.Read Appendix C.
NORMAL/REVERSE
switch
2.Check RS232C cable.
adaptor not used
3.Check that the Autograv and
faulty RS232C cable
computer communications
wrong computer
parameters are the same.
settings or component
used
Autograv CG3
Section
REFERENCE INFORMATION
AUTOGRAV SPECIFICATIONS
CG-3: 0.005 milligal
CG-3M: 0.001 milligal
Minimum Operating
Range:
Residual Long-term
Drift:
Typical Repeatability
in field use:
Range of Automatic
Tilt correction:
Dimensions:
Weight:
Power Consumption:
5 W at +25C.
Operating
Temperature Range:
-40C to +45C.
Optionally high temp. to +55C.
Interval Between
Readings in CYCLING
Mode:
Standard Memory:
Reference
Reading Resolution:
Reference Information
Noise Rejection:
Displayed and
Recorded Data:
Digital Display:
Keypad Input:
Standard Accessories
STANDARD ACCESSORIES
Tripod:
Battery:
Battery Charger:
Carrying Bag:
Communications:
Includes a special RS-232C data transfer cable and adaptor. Used for communicating
with peripheral devices.
Display Heater:
Reference
OPTIONAL ACCESSORIES
Belt Battery Pack:
Chart Recorder
Cable:
External Power
Cable:
Reference Information
Backpack:
High Temperature
Option:
Fo operating up to +55C
Elevation Options:
Scintrex can provide a variety of devices for elevation and positional control.
Peripheral Devices:
APPLICATION SOFTWARE
Scintrex supplies fully documented software written for the IBM family of
microcomputers, and certain other microcomputers which use the
MS-DOS operating system.
This software permits:
data reduction
archiving of data
processing of data
profile and contour plotting
modelling
TRAINING
Scintrex can provide detailed training programs for the CG-3 including operation, basic maintenance and data processing.
Description
Battery Pack Belt
Chart Recorder Cable
External Power Cable
Back Pack
Part Number
858 093
780 542
858 078
858 021
858 028
140 260
858 079
780 031
OPTIONS
Reference
Description
CG-3 Gravity Meter (includes:)
Case
Sensor
Tripod
Cover
Battery Charger
Console
Transport Case
Carrying Strap
Manuals
RS-232C Cable
RS-232C Cable Adaptor
RS-232C Female-Female
Gender Converter
Minor Spare Parts Kit
Carrying Bag
Battery
Display Heater Kit
Reference Information
Quantity
2
2
2
1
1
1
1
Part
Number
512 020
512 039
512 041
540 062
540 063
540 064
540 065
WARRANTY
AND
REPAIR
WARRANTY
All Scintrex equipment, with the exception of consumable items, is warranted against defects in materials and workmanship for a period of one year
from the date of shipment from our plant. Should any defects become evident under normal use during this warranty period, Scintrex will make the
necessary repairs free of charge.
This warranty does not cover damage due to misuse or accident and may
be voided if the instrument console is opened or tampered with by persons
not authorized by Scintrex Limited.
To validate the warranty, the warranty card supplied with the instrument
must be returned to Scintrex within 30 days of shipment from our plant.
Reference
REPAIR
When to Ship the Unit
Please do not ship your instrument for repair until you have communicated
the nature of the problem to our Customer Service Department by facsimile,
telephone or letter, etc. Our Customer Service Department may suggest certain simple tests or steps for you to do which may solve your problem without shipping the instrument back for repair. If the problem cannot be
resolved, our personnel will request that you send the instrument to our
plant for necessary repairs.
Reference Information
Description of Problem
When you describe the problem, include the following information:
the symptoms of the problem
how the problem started
if it is constantly present, intermittent or repeatable
if constant, under what conditions it occurs
a printout of measurement data demonstrating the problem
the initial constants that were used with the instrument
Please mention the instrument serial number in all communications regarding equipment leased or purchased from Scintrex.
Autograv CG3
Appendix
INTRODUCTION
The sensing element of the Autograv is based on a fused quartz elastic
system. The gravitational force on the proof mass is balanced by a spring
and a relatively small electrostatic restoring force. The position of the
mass, which is sensed by a capacitative displacement transducer, is altered by a change in gravity. An automatic feedback circuit applies DC
voltage to the capacitor plates producing an electrostatic force on the
mass which brings it back to a null position. The feedback voltage, which
is a measure of the relative value of gravity at the reading site, is converted
to a digital signal and then transmitted to the instrument's data acquisition system for processing, display and storage. The position of the center
of gravity of the proof mass relative to the case of the perimeter is shown
in Figure A1 on page A-2.
The inherent strength and excellent elastic properties of fused quartz together with limit stops around the proof mass permit the instrument to be
operated without clamping. Further protection is provided by a durable
shock mount system which attaches the sensor to the housing.
The parameters of the gravity sensor and its electronic circuits are chosen
so that the feedback voltage covers a range of over 7000 mGals without
resetting. The use of a low-noise electronic design, together with a highly
accurate auto-calibrating analog to digital converter, results in a resolution of either 0.005 or 0.001 mGal, equipping the gravity meter for both
detailed field investigations and large scale regional or geodetic surveys.
Theory
THEORY OF OPERATION
Theory of Operation
The instruments tilt sensors are also electronic, with a resolution of 1 arc
second. The outputs from the sensors are displayed on high resolution
meters on the instruments front panel and also transmitted to the data acquisition system where they are displayed and stored. If the instrument is operated on an unstable base, real-time corrections for tilt errors can be
automatically made over a range of 200 arc seconds.
Introduction
ever, since there is no critical operating point for the sensor, the upper operating temperature can be set at a lower or higher value (optional +35C or
+ 55C).
The sensor design is mechanically very simple. The fine balancing required
to obtain astatisation is not needed, as the displacement transducer has sufficient resolution (0.02nm) to detect the beam position of a non-astatised
system, and electronic filtering reduces the effect of seismic noise. The
mechanisms, micrometer screws, gearboxes and mechanical feed-throughs
associated with mechanical feedback systems have been replaced by a voltage applied to the same plates which form the displacement transducer. The
temperature control is also accurate enough for the sensor to operate without mechanical compensation.
Theory
Theory of Operation
Figure A2
A-4 Autograv CG3 Manual
Introduction
Console
functions
It processes and applies corrections to the signals from the sensor, stores
data, formats it for outputting and performs instrument control functions.
A menu format with prompts is used to operate the instrument.
Operating
modes
Console
operation
The control console operation is based on a one second gating time. Once
every second:
analog signals are sampled
the display is updated
keypad commands are responded to
and necessary control functions performed.
During a reading:
the gravity signal Vfb is sampled once every second by ADC1 (see
Figure A2 on page A-4).
The individual samples are averaged to filter out seismic noise.
A running mean of the samples is displayed together with its standard
deviation.
In the course of a reading, the input of ADC1 is periodically switched
to the calibration voltage (Vcal). The frequency of calibration depends
on the operating conditions and is determined by the operator before
the reading starts.
Theory
Theory of Operation
AND
(A.1)
The average value of SUmin is about 600 mGal. Depending on the values
GCAL1 and GCAL2 this value may differ somewhat for your meter. It is recommended that you calculate this value for your meter, because the gravity
readings of lower values than SUmin may not be accurate.
The high end of the gravity measurement range may be calculated by inserting value (SF/SC) = l.457 into Equation (A.4)
(A.2)
The average value of SUmax is about 8800 mGal. Depending on the values
GCAL1 and GCAL2 this value may differ somewhat for your meter. It is recommended that you calculate this value for your meter, because the gravity
readings of the higher value than SUmax may not be accurate.
The CG-3 sensor spring position is adjusted during the sensor manufacture
in such a way that, on the average, the gravity reading in Toronto (latitude
43.7 degrees) is about 4200 mGal. This, in conjunction with the gravity reading range of about 8000 mGals, assures that newly produced CG-3 may measure the relative gravity anywhere on the surface of the earth, from
elevations of a few thousand meters above the equator to sea levels at both
poles. For your convenience, values of normal gravity at sea level, as a function of latitude, are listed in Table I on page A-9.
Autograv CG3 Manual A-7
Theory
Theory of Operation
Because of the positive drift rate of the CG-3 gravity sensor, the uncorrected
gravity readings at the fixed locations gradually increase in value. The linear
drift of the CG-3 sensor is caused by an unavoidable creep of the quartz
spring, whose length under tension increases, on the average, by 0.5 ppm/
day at the operating temperature of 60C. A typical sensor with the drift rate
of 0.5 mGal/day will reach the high end of the measurement range, at the
latitude of 60 degrees, in about 20 years. If used at lower latitudes, as most
gravimeters are, the typical CG-3 service life will be even longer.
You may estimate the remaining service life of your meter in the following
way:
Example:
a.
b.
c.
If you are planning to use your meter at higher latitudes than your
present latitude, evaluate from Table I the gravity difference
between your present latitude and the highest latitude at which
you will be using your meter. Add this difference to the reading
obtained in Step b. to obtain the value SUm.
d.
1.
You have calculated for your meter from Equation (A.2) on page A-7:
SUmax = 8800 mGal
2.
3.
4.
5.
6.
7.
where:
gravity at latitude
(A.3)
Lat.
g
(Gal)
g - ge
(mGal)
Lat.
g
(Gal)
g - ge
(mGal)
Lat.
g
(Gal)
g - ge
(mGal)
0
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19
20
21
22
23
24
25
26
27
28
29
30
978.032680
978.034253
978.038969
978.046823
978.057805
978.071902
978.089098
978.109371
978.132697
978.159048
978.388392
978.220695
978.255916
978.294015
978.334944
978.378655
978.425095
978.474208
978.525935
978.580214
978.636978
978.696160
978.757689
978.821489
978.887485
978.955596
979.025740
979.097832
979.171785
979.247510
979.324915
0.000
1.573
6.289
14.143
25.125
39.222
56.418
76.691
100.017
126.368
155.712
188.015
223.236
261.335
302.264
345.975
392.415
441.528
493.255
547.534
604.298
663.480
725.009
788.809
854.805
922.916
993.060
1065.152
1139.305
1214.830
1292.235
30
31
32
33
34
35
36
37
38
39
40
41
42
43
44
45
46
47
48
49
50
51
52
53
54
55
56
57
58
59
60
979.324915
979.403907
979.484390
979.566266
979.649436
979.733799
979.839254
979.905696
979.993021
980.081122
980.169892
980.259224
980.349009
980.439138
980.529500
980.619987
980.710488
980.800892
980.891089
980.980970
981.070424
981.159343
981.247618
981.335140
981.421804
981.507502
981.592131
981.675586
981.757765
981.838568
981.917895
1292.235
1371.227
1451.710
1533.586
1616.756
1701.119
1786.574
1873.016
1960.341
2048.442
2137.212
2226.544
2316.329
2406.458
2496.820
2587.307
2677.808
2768.212
2858.409
2948.290
3037.744
3126.663
3214.938
3302.460
3389.124
3474.822
3559.451
3642.906
3725.085
3805.888
3885.215
60
61
62
63
64
65
66
67
68
69
70
71
72
73
74
75
76
77
78
79
80
81
82
83
84
85
86
87
88
89
90
981.917895
981.995650
982.071737
982.146062
982.218535
982.289066
982.357569
982.423960
982.488156
982.550080
982.609654
982.666806
982.721465
982.773564
982.823038
982.869828
982.913875
982.955125
982.993526
983.029033
983.061600
983.091188
983.117761
983.141285
983.161732
983.179075
983.193295
983.204373
983.212296
983.217053
983.218640
3885.215
3962.970
4039.057
4113.382
4185.855
4256.386
4324.889
4391.280
4455.476
4517.400
4576.914
4634.126
4688.785
4740.884
4790.358
4837.148
4881.195
4922.445
4960.846
4996.353
5028.920
5058.508
5085.081
5108.605
5129.052
5146.395
5160.615
5171.693
5179.616
5184.373
5185.960
Theory
Theory of Operation
If, after many years of service, your CG-3 has drifted out of range, then the
spring position may be reset by a routine procedure at the Scintrex plant.
You may also use this occasion to inquire about the available program to
overhaul your meter, to upgrade to the latest firmware version or to upgrade
to the CG-3M accuracy level (if the evaluation at Scintrex shows that the
quality of your sensor warrants such an upgrade).
PROCESSING
OF THE
GRAVITY SIGNAL
While reading this section, please also refer to the Signal Flow diagram in
Figure A2 on page A-4.
The input to the gravity analog to digital converter (ADC1) is switched between Vfb, the gravity transducer output and Vcal, the stable calibration voltage. When the instrument is in the ADJUST mode (after the START key is
pressed once), the calibration voltage is connected to ADC1. In this mode,
the value of Vcal can be viewed on the display under the heading GRAV.
In the READ mode, (after the START key is pressed a second time) the input
to ADC1 is held at Vcal for one second (during PAUSE) then switched to
Vfb for x one second samples. It is then switched back to Vcal for one sample then to Vfb for x samples and so on until the reading is terminated. The
value of x is entered under the Autograv setup at the CAL AFTER prompt.
The output of ADC1 is termed SC when Vcal is applied and SF when the Vfb
is applied.
Uncorrected
sample
The calibration voltage and other calibration factors are then applied to the
gravity signal by the function fcal in order to give an uncorrected sample
SF
SF 2
SU = f cal(SC, SF) = GCAL1 ------- + GCAL2 -------
SC
SC
(A.4)
SF is divided by the most recent value of SC to correct for any drift in the
scale factor of ADC1. The factors GCAL1 and GCAL2 are the instrument
calibration factors which are entered via the keypad. They take into account
a small quadratic non-linearity inherent in the ADC and in the conversion
from electrostatic feedback voltage to force.
The uncorrected sample has drift correction and GREF applied and if selected, then also the tilt correction, to give a corrected sample
(A.5)
S = SU DC + GREF + TIC = SU + C
SF ij
SF ij 2
GCAL1 ---------- + GCAL2 ---------- + C ij
SC j
SC j
(A.6)
( Sij + Cij )
j = 1i = 1
= 1i = 1
= j--------------------------------------------RU ij = ---------------------------------------------------------------------------------------------------------------------( j 1 )x + i NR
DUR
Where SCj is the value of the jth voltage calibration and SFij and Cij are the
ith gravity sample and correction after the jth voltage calibration.
Standard
deviation
(A.7)
where NR is the number of rejections. On the right hand side of the second
line the display alternates between the latest value of the SD during calibration sampling and the standard deviation of the mean, which is estimated
under the assumption that the noise is normally distributed (i.e. white) from
the relationship:
SD
ERR = ----------------DUR
(A.8)
Theory
Theory of Operation
The formula given in Equation (A.8) is the one used in the CG-3. However,
during periods of high seismicity (caused by microseismic noise), the standard deviation of the mean is better estimated by
SD
ERR = ------------DUR
(A.9)
Sampling of the gravity signal stops when either DUR reaches the preset
READ TIME, when ERR < ERR LIM or when the START/STOP key is
pressed. Up to three further corrections: the tilt correction (TIC), the temperature correction (TEC) and the earth tide correction (ETC) are then
added to RU to give the reading R which is then displayed.
(A.10)
COMPENSATION
AND
CORRECTIONS
DRIFT CORRECTION
Long term instrument drift is largely due to stress relaxation in the elastic system. After an initial stabilization period, it can be considered a linear function of time. The effect of long term drift will be to produce a uniform
change in reading with time. For example, Figure A3 on page A-17 shows
several days of continuous readings. If two uncorrected readings are taken
at times t1 and t2* at a point where there has been no change in gravity, then
the difference between them will depend on the long term drift rate, d, of
the instrument and is:
(A.11)
(A.12)
DRIFT is the drift constant with units of mGal/24hrs, ts is the drift correction
start time and t is the time at which the samples are taken. Both DRIFT and
ts are entered via the keypad.
With the drift correction applied, Equation (A.11) becomes:
R ( t 2 ) R ( t 1 ) = RU ( t 2 ) DC ( t 2 ) RUt 1 + DC ( t 1 )
= ( d ( t 2 t 1 ) DRIFT(t 2 t 1) )
(A.13)
(A.14)
R(t 2) R(t 1) = 0
The procedure for adjusting the drift correction is given in the Getting
Started section starting on page 2-41.
(A.15)
g is the value of gravity at the reading site and x and y are the tilts of the
gravity sensor about two perpendicular horizontal axes (x and y) as shown
in Figure 8 on page 2-32 with x = y = 0 being defined as the orientation in
which the uncompensated gravity meter reading is maximized.
* calculation precision limits the maximum value of (t-ts) to 4 months (128 days)
Theory
Theory of Operation
The tilt correction operates over a range of 200 arcsec and is
TIC = g t ( 1 cos x cos y )
(A.16)
where gt is an average sea level gravity value of 980.6 Gal and X and Y are
the indicated gravity meter tilts.
Assuming gt = g (this assumption leads to an error of 0.002 mGal in the worst
case), the corrected reading will be:
R( x, y) = RU(0, 0) g t ( cos x cos y cos x cos y )
(A.17)
If
x = x
y = y
(A.18)
then
R( x, y) = RU(0, 0) = R(0, 0)
(A.19)
and errors due to instrument tilt are eliminated. The method used to adjust
X and Y so that Equation (A.18) is satisfied, is given in the Maintenance section starting on page 6-4.
There are two software selectable modes of operating the tilt correction.
1.
2.
The reason for two modes of operation is that the tilt sensors have a relatively
fast response (typically a few seconds) up to within approximately 5 arcsec
of the final value and then it takes up to a minute to come to within one arcsec of the final value.
If the Autograv is on a solid base and it does not move during the reading,
then the DISABLE mode will give the most accurate result as the tilt sensors
will have more time to settle before a compensation is applied.
(A.20)
TEMPCO is the instrument temperature coefficient in mGal/mK. It is measured during the production of each instrument and is entered through the
keypad. A new value of TEC is computed and applied after the last sample
has been taken.
The spring temperature TEMP in units of mK can be displayed on the liquid
crystal display. The last TEMP value for each reading is also stored in memory together with other measurement variables. The range of TEMP is
3.0 mK, and it is adjusted with a zero offset to within 1.0 mK for normal
operation (See Getting Started section Temperature Compensation on
page 2-35). This ensures that there is at least a 2.0 mK useful range to compensate for severe ambient temperature fluctuations.
Theory
It should be noted that over a period of several hours, some tilt changes may
arise from movement of the elastomer in the shock isolators (used to protect
the sensor from severe shocks). This not relevant for the FIELD mode. In the
CYCLE mode, however, these changes are corrected by the tilt correction
function, regardless of whether or not the tilt correction function is enabled
or disabled.
Theory of Operation
The chamber temperature is measured by a set of two thermistors, which
themselves exhibit a long-term drift due to their aging. However, the effect
of this drift is indistinguishable from the gravity sensor drift, and is corrected
by the linear drift correction.
Residuals
The ETC will compensate for the gravity effects of the Sun and Moon to within (usually) 3 Gals. Tidal accelerations cause elastic deformations to the
Earths body (earth body tides). The average effect of this deformation is taken into account by the gravimetric factor 1.16 in the Longman formula (16%
increase over the amplitude of gravimetric tides of the rigid Earth). Slight
variances of gravimetric factor exist. The gravimetric factor is latitude dependent and different tidal waves have slightly different factors. This may result
in minor residuals of a semi-diurnal and diurnal nature having amplitudes
of about 3 Gals. In addition, the ocean tides, besides their direct gravitation, cause additional deformation by their loading. These deformations
lead to additional changes of the gravity values observed on the surface of
the Earth. The residuals from both effects could be effectively removed from
field measurements by using another gravity meter as a base-station (Figure
A3 on page A-17).
* Longman, I.M., Journal of Geophysical Research, Volume 64, No. 12, Formulas for Computing
the Tidal Accelerations Due to the Moon and the Sun, December 1959
Theory
Figure A3Continuous measurement over ten days showing the
linear nature of drift together with other phenomena,
such as ocean loading tide.
Theory of Operation
Autograv CG3
Appendix
Hysteresis
Hysteresis
Figure B2 Retardation effect after unclamping LaCoste Romberg
gravimeter G79, previous clamping time 20 minutes.
EFE Hanover (from Gravimetry by Wolgang Torge,
Walter de Gruyter, 1989.)
Autograv CG3
Appendix
INTERFACING WITH
THE AUTOGRAV
RS 232C STANDARD
The standard defines 22 pins and 3 unassigned pins. This does not imply a
25 pin connector, although this has become the industry standard. Conventionally, the male connector is used with the DTE and the female connector
with DCE.
The Autograv used the IGS console as the data acquisition and control system. The IGS is RS 232C compatible. This does not mean it can immediately
communicate with every other device so labelled. The statement RS 232C
compatible means that none of the specifications defined by the standard
EIA RS 232C are exceeded by the equipment. There may exist minor differences within this specification that preclude immediate communication.
These incompatibilities may generally be overcome by reviewing pin assignments and re-wiring (by means of adapters) as necessary.
Autograv CG3 Manual C-1
Interfacing
The Electronic Industry Association (EIA) established a standard for the interface between data terminal equipment (DTE) and data communication
equipment (DCE) in 1963. The latest revision of this standard has been in effect since 1969, and is known as RS 232C. This standard specifies mechanical and electrical characteristics of the interface, the number of interchange
circuits and their functions, and the relationship between these circuits and
standard interface types. Standards similar to the RS 232C have also been established by the Comit Consultatif International Tlphonique et
Tlgraphique (CCITT). These standards, CCITT V.24 and CCITT V.28, are
very similar to RS 232C but not identical.
IGS OUTPUT
SERIAL I/O CONNECTIONS
The IGS is configured as Data Communications Equipment (DCE). Communication with other devices is via the RS-232C DB25S (female) connector on
the front of the Autograv. Supplied with the instrument are a bi-directional
serial RS 232C* cable with a NORMAL/REVERSE switch, and an IGS-2
adapter.
The adapter is first plugged into the instrument, then the cable (either end).
The function of the NORMAL/REVERSE switch is to allow the instrument
to be reconfigured as Data Terminal Equipment (DTE). This is performed by
switching to the REVERSE position. This is necessary when attempting to
communicate with another device configured as DCE. When the switch is
set to NORMAL the instrument remains DCE. Generally, the NORMAL position is used to communicate with printers and computers. Try the REVERSE if the NORMAL does not work.
This switching is necessary because of the assignment of functions to certain pins in the RS 232C standard. The pins in question are 2 and 3. The DCE
mode expects to receive data on pin 2 and transmit data on pin 3 and the
assignments are compatible for DTE communication. That is, DTE transmits
through pin 2 and receives data on pin 3.
When interfacing two pieces of DCE, it is necessary to switch to REVERSE
so the signals are received on the appropriate pins (see Figure C1).
..
DCE
Function
DTE
to
Pin
Pin
Receive
Receive
Transmit
Transmit
Function
IGS Output
DCE
Function
DTE
to
Pin
Pin
Function
REVERSE
DCE
Function
DTE
to
Pin
Pin
Function
Receive
Receive
Transmit
Transmit
NORMAL
Figure C1 Function of the NORMAL/REVERSE Switch
on the Serial RS-232C Cable
Receive
Transmit
Ground
14
EIAR1 (DTR)
18
Positive Battery
24
D/A out
The adapter between the Autograv and the serial RS 232C cable must be
used. Damage may occur to either or both instruments if the adapter is not
in place. The adapter wire as indicated in Figure C2 and the serial RS 232C
cable wiring appears as Figure C3.
Interfacing
Pin
Female
(Into Serial RS-232C
Cable)
Male
(Into Autograv)
Description
Pin
Pin
Receive
Transmit
Transmit
Receive
18
Clear to Send
Carrier On
Positive Battery
Description
Ground
Signal Ground
D/A out
24
25
Analog Chart
Recorder
EIAR1
14
20
Data Terminal
Ready (DTR)
Figure C2
IGS-RS 232C Wiring Diagram (P/N 780 541)
NORMAL
REVERSE
Male
Male
Transmit
Receive
Request to Send
IGS Output
Male
NORMAL
Male
5
Clear to Send
Ground
Carrier On
20
20
REVERSE
Interfacing
Baud rateData are transmitted serially at a selectable baud rate between 110 and 2400 baud.
COMMUNICATION SOFTWARE
Once the Autograv has been physically linked to a computer by means of
adapters and serial cables, it is necessary to make them communicate. The
interface software may be a commercially available communications
package, or may be custom software dedicated to the users hardware
configuration.
Commercial communications packages tend to be quite comprehensive
and for the simple function of transferring data to a computer from the
Autograv are unnecessarily complicated.
Scintrex provides for the Autograv users communication software called
IDUMP. This program is the property of Geosoft and is provided free of
charge. Please observe all Geosoft requirements in the use and distribution
of this software as stated in the copyright message. It supports serial port
COM1/COM2 and a baud rate of up to 2400 BPS (for Autograv). It is self-documented and easy to use. Should you have any questions or concerns,
please contact Scintrex Ltd.
Autograv CG3
Appendix
Code
Value
(GCAL1)
Gravity constant #2
(GCAL2)
Temperature constant
(TEMP)
Tilt X sensitivity
(TILTXS)
Tilt Y sensitivity
(TILTYS)
Drift constant
(DRIFT)
Constants
Gravity constant #1
Autograv Technician
Autograv CG3
Index
Index
A
Index
Accessories
Optional 83
Part numbers 85
Standard 82
ADJUST mode 225
Adjusting
Drift Correction 239
Temp. compensation offset 234
Tilt correction 64
Tilt sensor
coarse zero 610
Cross-coupling 617
sensitivity 612
zero 64
X-axis Sensitivity (TILTXS) 613
Y-axis Sensitivity (TILTYS) 615
Auto Rejection 44
Autograv
Handling 46
Incomplete warmup stabilization 410
Initialization constants
losing 52
Instrument parameters 31
Interfacing with C1
Keypad 26
Leveling 228
Maintenance 61
Optional Accessories 83
Overview 11
Powering up 24
Reset 72
Setup 218
Specifications 81
stabilization period 23
Standard Accessories 82
start-up sequence 23
Storing 620
Transporting 46
Troubleshooting 71
unpacking 22
warm-up period 23
Autograv Setup
AUTO REJECTION 322
AUTO-RECORD 325
CAL. AFTER 321
CONT. TILT CORR. 321
CYCLE TIME 317
DRIFT CORRECTION 326
INITIAL CONSTANTS 328
MODE 316, 317
READ TIME 317
STOP ERR. <LIM> 320
Submenus 316
TIDE CORRECTION 322
AUTOGRAV START DUMP 331
Automatic
Record 325
Rejection 322
Shutoff 32
Station increment 39
Aux (Auxiliary) key 28
Aux. mode
locking 340
unlocking 340
Auxiliary mode 31, 32
Flowchart 214
B
Battery 52
Changing 58
Charging time 56
Checking voltage 53
Cold weather operation 410
Compartment Fuse 58
External belt 52
Life expectancy 54
Low warning 53
Panasonic
12V/7.2Ah 52, 55
LCR6V/7.2P 55
Replacing 55
Sonnenschein Dryfit A200 12V/6Ah 55
Temperature effects 52
voltage display update 54
BAUD RATE 336
BIAS1 333
BIAS2 335
Bottom function keys 29
C
CAL. AFTER 321, 42
Calculating
Drift values
example 239
Cancelling
measurement 318
Change key 28
Changing
Battery 58
Start time 327
Charger 55
fuse 24, 57
Selecting alternate voltage 56
Charging time 56
Chart Recorder 63
CHART SEL 319
Index-ii Autograv CG3 Manual
Checking
Battery voltage 53
Temperature compensation 231
CHG? prompt 212
Clear key 29
Cold Weather Operation 410
Communication
Software C6
IDUMP 411, C6
Compensation
Temperature 61
Checking 231
Console
Operation A5
CONT. TILT CORR. 321, 43
disabled 43
Selected 43
Control Console A5
Correction
Drift 236, 62, A12
Tide A16
Tilt A13
Continuous 43
CR DELAY 338
CYCLE TIME 317
CHART SEL 319
minimum value 318
CYCLING
Mode 317
D
Data
Auto rejection 44
Displayed 82
dump 411
cannot 72
Entry
submenus 32
Key 27
E
Earth Tide Correction (ETC) A16
Earthquake shock waves 48
EDIT mode 339
Eliminating
noise spikes 44
Enter
Key 29
Mode 26
Keys 29
ENTER? prompt 212
EPROM chips 31
ERASE MEMORY 34
Erasing memory 217
Error display 225
Error limit stop 320
ETC A16
Examples
Calculating drift values 239
Changing read time 212
Data output printout 237
Initialise submenus 33
moving between menus 211
selecting a menu 211
Exiting
Memory Mode 246
Extender legs
Tripod 229
External battery belt 52
F
Field
Measurements
Mode 317
setting-up 241
Mode
Data recall 242
Operation 41
FLD GRAV 317
Autograv CG3 Manual Index-iii
Index
Notebook feature 45
Output 237
Recall
Cycling mode 246
Field mode 242
recalling 242
Recording 44
Serial format C5
Types recorded 82
DATA BITS 337
Date
Reset 215
DCE C2
Decimal key 29
DEG. LAT 323
DEG. LONG 324
Digital display 82
Direction key 29
Display 210
battery voltage update 54
digital 82
error 225
gravity reading 225
Heater
Minor spare parts kit 83
Leveling parameters 225
Parameter set-up 210
SD 225
Drift
calculating example 239
Correction 236, 62, A12
adjusting 239
obtaining accurate 62
setting 326
measuring 236
residual long-term 81
DRIFT CORRECTION 326
DTE C2
DUMP ALL DATA 331
Dumping Data 411
Dust cover 46
removing 22
G
Flowchart
Auxiliary Mode 214
Format
Serial data C5
Fuse
Battery compartment 58
Charger 57
charger 24
G
GCAL1 618, A7
GCAL2 618, A7
Geosoft 411, C6
GMT. DIFF 324
Gravimeter
recalibrating 618
Gravimetric factor A16
Gravity
Stored readings A6
Value display 225
Gravity offset 325, A17
Gravity signal
Processing A10
Greenwich Mean Time 324
GREF A11, A17
GRID 36
Ground conditions 48
H
Handling 46
rough 47
HEATER 312
Hints
Cold weather operation 410
Earthquake shock waves 48
industrial traffic noise 48
Microseismic noise 48
Index-iv Autograv CG3 Manual
Operation 46
poor ground conditions 48
Transportation 47
Wind noise 49
I
IDUMP 411, C6
IGS Output C2
Industrial traffic noise 48
Info key 28
INITIAL CONSTANTS 328
Memory loss 52
INITIALISE
Menu 33
Submenus 33
Initialise
AUTO STN INC 39
ERASE MEMORY 34
GRID 36
HEATER 312
JOB NO. 35
LANG. ENGLISH 312
LINE 37
LINE DIR 36
OPERATOR 35
REC AUTOSTART 311
SERIAL # 311
STATION 38
TIME 313
Initializing 215
Parameters/Software 215
Instrument parameters 31
Interfacing
Autograv C1
J
JOB NO. 35
L
LANG. ENGLISH 312
Latitude 323
Leveling 228
Parameters display 225
Requirements 43
Life expectancy
Batteries 54
LINE 37
M
Main Menu
(Aux. Mode) 31
flowchart 214
Maintenance
Autograv 61
Tripod 620
Measurement
cancelling 318
Field setup 241
Measuring
Drift 236
gravity 241
Tilt Sensor cross-coupling 616
X-axis Sensitivity (TILTXS) 612
XE 64
Y-axis Sensitivity (TILTYS) 614
YE 68
Memory
Directly accessing records 245
erasing 217
keys 27
Loss 52, 72
Mode
Exiting 246
Scrolling 242
Autograv CG3 Manual Index-v
Index
Line
direction 36, 310
key 27
separation 314
LINE DIR 36
LINE SEP 314
Line Separation mode 314
LOCK AUX. 340
Locking Auxiliary mode 340
Longitude 324
Longman formula A16
Low battery warning 53
N
Standard 81
Menu
INITIALISE 33
Main 31
flowchart 214
Menus
moving between 210
Selecting 211
Structure 210
Method key 27
Microseismic noise 48
Minimum cycle time 318
MODE
Selection 317
Mode
Adjust 225
AUTOGRAV SETUP 316
Aux.
Locking 340
unlocking 340
Auxiliary 32
Auxiliary (Main Menu) 31
CYCLING 317
recalling data 246
EDIT 339
ENTER 26
Field
data recall 242
FLD GRV 317
LINE SEP 314
Memory
Exiting 246
NONE 317
OPERATION 26
OUTPUT 330
Pause 225
STN SEP 315
Modes
operating 26, A5
Modifying
parameters
Index-vi Autograv CG3 Manual
CHG? 212
paramters 212
Motion Noise 48
N
NE/SW key 29
Noise A11
Earthquake shock waves 48
Industrial traffic 48
Microseismic 48
moition 48
Rejection 82
spikes
eliminating 44
Wind 49
NONE 317
Notebook Feature 45
Numeric keys 29
O
On/Off key 27
Operating
modes 26, A5
Operation
Cold weather 410
Console A5
Hints 46
Keypad 41
Mode 26
Keys 27
OPERATOR 35
Options
Part numbers 85
Output 332
AUTOGRAV START DUMP 331
BAUD RATE 336
CR DELAY 338
R
Data 237
DATA BITS 337
DUMP ALL DATA 331
IGS C2
Mode 330
Submenus 331
Index
P
Panasonic
12V/7.2Ah battery 52
LCR6V/7.2P 55
Parameters
Initializing 215
Instrument 31
modifying 212
ENTER? 212
SELECT/ENTER 212
Parameters set-up
display 210
Parts List 85
Pause mode 225
PLOT 332
parameters 332, 334
BIAS! 333
BIAS2 335
SCALE2 335
START PLOT 336
TRACE1 332
TRACE2 335
Trace 332
Plus/Minus keys 29
Power
Consumption 81
loss 51
Off
Consequences 51
Off Time and stabilization 52
Powering up 24
Processing
Q
Quartz elastic system A1
R
Range
Automatic Tilt correction 81
Operating temperature 81
Readings A7
Ratio
Internal Cal. to Gravity samples 42
READ TIME 317
Reading
Out-of-range 71
Range A7
Repeatability 81
Resolution 81
Terminating 42
Time 42
Real Time Clock 82
REC AUTOSTART 311
Recalibrating gravimeter 618
Recalling Data
Cycling mode 246
Field mode 242
Record key 27
Recording
Data 44
Notebook feature 45
Records
Autograv CG3 Manual Index-vii
S
Scanning 243
Repair 87
Repeatability
readings 81
Replacing Batteries 55
Reset
Autograv 72
date 215
time 215
Resolution
Reading 81
RS 232C C1
S
SCALE1 334
SCALE2 335
Scanning
Records 243
Scroll keys 28
Scrolling in memory 242
SD display 225
SELECT/ENTER prompt 212
Selecting
Charger voltage 56
Mode of operation 317
Sensor
Latched 71
Unlatching 71
SERIAL # 311
Serial data format C5
Service Life
Spring A7
Setting
Drift correction 326
Setting-up
Autograv 218
Field Measurements 241
Tripod 226
Sign conventions 244
Index-viii Autograv CG3 Manual
Software 215, 84
Communication C6
Sonnenschein Dryfit A200 12V/6Ah 55
Spare Parts Kit 86
Springs
Service Life A7
Stress relaxation B1
Stabilization period 23
Standard deviation A11
value display 225
START PLOT 336
Start/Stop key 29
Station
Auto-increment 39
disabled 310
seleceted 39
Initialise 38
Key 28
Separation 315
Station Separation mode 315
STOP ERR. 320
Storing
Autograv 620
Stress relaxation
Springs B1
Submenus
Autograv Setup 316
Data entry 32
Initialise 33
Output 331
Returning from 32
T
TEC A15
Temperature
Compensation 61, A15
adjusting offset 234
checking 231
offset 232
W
Cycle 317
Initialising 313
reading 42
Reset 215
Top keys 27
Trace 332
TRACE1 332
TRACE2 335
Training 84
Transportation 47
Autograv 46
case 22
Rough handling 47
Tripod
Extender Legs 229
Maintenance 620
on a hard surface 43
on a soft surface 229
setup 226
Troubleshooting 71
U
Unlatching sensor 71
Unlocking Auxiliary mode 340
Unpacking
Autograv 22
V
Vee-groove plate 617
Voltage 53
checking 53
W
Warm-up period 23
Warning
Autograv CG3 Manual Index-ix
Index
TEC A15
Effects
Batteries 52
Terminating
Readings 42
Text transmission sequence C6
Theory of Operation A1
Thermostat
stability 51
TIC A13
Tide
Correction A16
TIDE CORRECTION 322
DEG. LAT 323
DEG. LONG 324
GMT. DIFF 324
Tilt
Correction A13
Adjustment 64
continuous 43
TIC A13
Tilt Sensor
Adjusting
cross-coupling 617
sensitivity 612
zero 64
Coarse Zero Adjustment 610
Cross-Coupling 616
adjusting 617
measuring 616
Sensitivity Adjustment (tiltxs, tiltys) 612
Zero Adjustments 64
TILTXS 612
adjusting 613
measuring 612
TILTYS
adjusting 615
measuring 614
Time
Battery charging 56
Changing start 327
X
Low battery 53
Warranty 87
Weight 81
Wind 49
X
XE
Coarse adjustment 610
Fine Adjustment 66
Measuring 64
Y
YE
Coarse adjustment 611
Fine adjustment 69
Measuring 68