GE Dash 2500
GE Dash 2500
GE Dash 2500
GE Healthcare
NOTE: In addition to software version RAJ, the information in this manual also applies to the old Dash 2500 Patient
Monitor software version . There are no user-apparent differences among these software versions. Due to continuing
product innovation, specifications in this manual are subject to change without notice.
NOTE: For technical documentation purposes, the abbreviation GE is used for the legal entity name, GE Medical
Systems Information Technologies.
Listed below are GE Medical Systems Information Technologies trademarks. All other trademarks contained herein are the
property of their respective owners.
Ohmeda Oximetry and other trademarks (OxyTip+, PIr, TruSat, TruSignal, TruTrak+) are the property of GE Medical Systems
Information Technologies, a division of General Electric Corporation. All other product and company names are the
property of their respective owners.
Alaris Turbo Temp is a trademark of ALARIS Medical Systems.
SuperSTAT, Trim Knob and EK-Pro are registered trademarks of GE Medical Systems Information Technologies.
MASIMO SET is a trademark of Masimo Corporation. Possession or purchase of this device does not convey any express or
implied license to use the device with replacement parts which would, alone, or in combination with this device, fall within
the scope of one or more of the patents relating to the device.
NELLCOR, OxiMax, C-LOCK and SatSeconds are trademarks of Nellcor Puritan Bennett.
T-2
2042481-001 C
Jan. 2009
Contents
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-1
Manual Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-3
Revision History . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-3
Manual Purpose . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-3
Intended Audience . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-3
Ordering Manuals . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-3
Illustrations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-4
Safety Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-4
Responsibility of the Manufacturer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-4
General . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-4
Warnings, Cautions, and Notes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-5
Equipment Symbols . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-6
Service Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-8
Service Requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-8
Equipment Identification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-8
3
Revision C
Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-1
Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-3
Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-1
Maintenance Schedule . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-3
Manufacturer Recommendations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-3
Manufacturer Responsibility . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-3
Cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-4
Periodic Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-6
Integrity of Hoses and Cuffs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-6
Long-Term Storage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-6
Periodic Procedures . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-6
Care of Storage Batteries . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-7
Electrical Safety Tests . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-8
General . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-8
Test Equipment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-8
Safety Resistance Testing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-8
Patient (Source) Leakage Current Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-10
Patient (Sink) Leakage Current Testing
(Mains Voltage on the Applied Part) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-12
Checkout Procedures . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Parameter Level Functional Testing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Service Mode Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Service Mode Exit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
4-13
4-13
4-14
4-31
Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-1
Required Service Equipment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-3
General Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-4
Parameter Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-5
Error Codes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-6
Alarm Code Interpretation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-6
System Failures . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-6
ii
Revision C
6-14
6-14
6-14
6-15
6-15
Revision C
iii
B
C
Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .C-1
Monitor Configuration Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Enter Configuration Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Configure Default Changes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Exit Configuration Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
C-3
C-3
C-3
C-4
Connection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . F-1
Technical Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . F-3
Product Compatibility . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .F-3
External Connectors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . F-4
Ethernet Port . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .F-4
iv
Revision C
Revision C
vi
Revision C
Revision C
Introduction
1-1
1-2
Revision C
Manual Information
Revision History
Each page of this manual has the document part number and revision letter at
the bottom of the page. The revision letter identifies the documents update
level. The revision history of this document is summarized below.
Revision
Comment
ECO release
ECO release
Manual Purpose
This manual supplies technical information for service representatives and
technical personnel so they can maintain the equipment to the assembly level.
Use it as a guide for maintenance and electrical repairs considered field
repairable. Where necessary the manual identifies additional sources of relevant
information and technical assistance.
See the operators manual for the instructions necessary to operate the
equipment safely in accordance with its function and intended use.
Intended Audience
This manual is intended for service representatives and technical personnel who
maintain, troubleshoot, or repair this equipment.
Ordering Manuals
To order additional copies of this manual, call Accessories and Supplies and
request part number 2042481-001 for a paper copy, and 2042482-001 for a CD
copy. Refer to the How To Reach Us page for Accessories and Supplies contact
information.
WARNING
This Service Manual is available in English only except as
otherwise expressly required by local law or agreed to at a
local level.
If a customers service provider requires a language other
than English, it is the customers responsibility to provide
translation services.
Do not attempt to service the equipment unless this
Service Manual has been consulted and is understood.
Failure to heed this Warning may result in injury to the
Revision C
1-3
Illustrations
All illustrations in this manual are only examples, and may not necessarily
reflect your system settings or data displayed in your system. If a particular
selection is not available in your system, the selection is shown grayed in the
menu.
Safety Information
Responsibility of the Manufacturer
GE is responsible for the effects of safety, reliability, and performance only if:
General
This device is intended for use under the direct supervision of a licensed health
care practitioner.
This device is not intended for home use.
Federal law restricts this device to be sold by or on the order of a physician.
Contact GE for information before connecting any devices to the equipment that
are not recommended in this manual.
Parts and accessories used must meet the requirements of the applicable IEC
601 series safety standards, and/or the system configuration must meet the
requirements of the IEC 60601-1-1 medical electrical systems standard.
Periodically, and whenever the integrity of the device is in doubt, test all
functions.
The use of ACCESSORY equipment not complying with the equivalent safety
requirements of this equipment may lead to a reduced level of safety of the
resulting system. Consideration relating to the choice shall include:
If the installation of the equipment, in the USA, will use 240V rather than 120V,
the source must be a center-tapped, 240V, single-phase circuit.
1-4
Revision C
Revision C
1-5
Equipment Symbols
NOTE: Some symbols may not appear on all equipment.
ATTENTION: Consult accompanying documents before using the equipment.
Battery in use
Storage temperature
This side up
Keep dry
1-6
Revision C
Revision C
1-7
Service Information
Service Requirements
Follow the service requirements listed below.
Equipment Identification
Every GE device has a unique serial number for identification. A sample of the
information found on a serial number label is shown below.
### ## ## #### # #
Description
A
product code1
year manufactured
manufacturing site
miscellaneous characteristic
1. Current Dash monitor product code SCG
1-8
Revision C
Revision C
Equipment Overview
2-1
2-2
Revision C
Components
Equipment Description
The Dash 2500 patient monitor is intended to monitor and measure oscillometric
noninvasive blood pressure (systolic, diastolic, and mean blood pressure), heart
rate/pulse, respiration rate, ECG with lethal arrhythmia (VTACH, VFIB/VTACH, and
asystole), oxygen saturation (SpO2) by noninvasive pulse oximetry, and
predictive temperature with an electronic thermometer for adult, pediatric and
neonatal patients. The Dash 2500 Monitor also detects alarm limit conditions
and is capable of recording up to two waveforms. Using this Monitor, a clinician
can view, record, and recall clinical data derived from each parameter.
The Dash 2500 can be configured with or without a printer and/or temperature
module. Also, the units are ordered with either Masimo or Nellcor SpO2.
The network function is optional for Dash 2500.
Configurations
Each monitor is supplied with an accessory pack. The contents of the pack vary
according to model. Unpack the items carefully, and check them against the
checklists enclosed within the accessory boxes. If an accessory is missing or if
an item is in a nonworking condition, contact GE Medical Systems Information
Technologies immediately.
NOTE: Perform regular functional testing of each of the parameters, using the
accessories supplied with the Dash 2500 Monitor.
It is recommended that all the packaging be retained, in case the Monitor must
be returned for service in the future.
Revision C
2-3
Monitor
AC Power Indicator
The indicator lights green when AC mains power is applied to the Monitor. The
indicator does not illuminate when the Monitor has no AC mains power.
2-4
Revision C
Main - closes the menu system and takes you back to the main screen.
Revision C
2-5
Theory Of Operation
The Monitor is a portable unit that receives input power from an external AC
source, or internal rechargeable battery.
When the Power button is pressed, the Main Board is brought out of a sleep
mode and turns on the power regulators. The power regulators provide
conditioned power from one of the two input power sources:
AC Mains
Internal battery
SpO2 sensor
ECG leads
Cuff
Interface with a central station or other device is accomplished through the 9pin host communication port or the 15-pin wireless communication port on the
back of the Monitor.
2-6
Revision C
DINAMAP
Revision C
2-7
2-8
Revision C
Revision C
Function
Pin
Function
Ground
DCD
TX2_RS232
RX2_RS232
Fused +5 volts
DTR
GND
No connection
DSR
Ground
RTS
CTS
No Connection
No Connection
10
No Connection
11
12
13
14
No connection
15
No connection
2-9
Functional Description
The following section provides the functional interface relationship. The Monitor
contains a number of electrical and electro-mechanical assemblies. These
assemblies are:
PSU Module
DC2DC PWA
Main Board
UI PWAs
DAS PWA
SpO2 PWA
Interface PWA
Printer (optional)
NiMH Battery
PSU Module
The PSU module is an AC Mains to DC converter. The module receives AC power
from the mains input connector. When AC INPUT is applied to the module, the
module AC/DC Converter changes the AC INPUT supply via rectifier circuit to a
high voltage DC. The DC power is then routed through a high frequency
switching converter and regulated to 12 VDC. This supply is connected to the
DC2DC PWA for further regulation.
DC2DC PWA
The PSU supplies regulated DC power to Monitor. The DC2DC PWA is designed to
operate from the output of the AC MAINS PSU module
(+12 VDC), or from an internal NiMH rechargeable battery (+8.4 VDC). The PSU
automatically selects the power source based on the following priority:
The DC2DC PWA converts the selected power source into the following main
voltages:
2-10
+5V
+12V
+3.3V
Revision C
Main Board
The Main Board is configured with Flash ROM, EEPROM, RAM, Primary Processor,
Secondary Processor, and NIBP. The Primary Processor operates from a 4.9152
MHz crystal stepped up to 49.152 MHz. The Primary Processor services and
controls the RAM, Flash ROM, EEPROM, the physiological interface modular
devices and display backlighting. The Secondary Processor monitors the power
supply circuit and signals within the NIBP circuits, controls the power-on/off
sequences, and performs watchdog tasks on it and the Primary Processor
monitors. The Secondary Processor monitors the power supply circuit and
controls the battery back up enable when no external sources are present and
shuts down the unit when the battery is exhausted. It enables the battery
charging circuit based on the battery charge status, unit power consumption,
and the availability of an external power source.
The Random Access Memory (RAM) is composed of a SRAM chip and two SDRAM
chips. The 512 Kbytes of battery-backed SRAM is provided to store trend data
and to provide space for working algorithms and is accessed on bits D[0:15] of
the data bus. The two 64-Mbit SDRAM chips are set up to form a 32-bit data bus
on bits D[31:0] that is used for running the program and working memory. This
gives 16 Mbytes of memory with an access time of less that 20 ns. The program
is loaded (including the boot code) from the 16-bit FLASH Read Only Memory
(ROM). The Electronically Erasable Programmable Read Only Memory (EEPROM)
is an 8-bit chip that is used to store the calibration and other "setting" variables
that have to be maintained in the event of a complete power failure.
If a hardware or software error causes a malfunction, its watchdog will provide
an internal and external RESET (L) signal. The FAILSAFE controller causes the
FAILSAFE (L) signal to go low. This signal passes to the Secondary Processor,
which disables the Primary Processor's power supplies, thus turning it off.
FAILSAFE (L) also passes to the control logic, which dumps the cuff pressure. The
system is left in a safe state but remains ON to enable the Secondary Alarm to
stay active. The Primary Processor monitors the activity of the secondary via its
handshaking communications. If the Secondary fails, the Primary can assert the
FAILSAFE line by overriding the FAILSAFE controller. The Secondary Alarm is a
hardwired alarm that will sound in the event of a FAILSAFE condition. Pressing
Revision C
2-11
SpO2 PWA
The Monitor can be configured with either NELLCOR or MASIMO SET SpO2
hardware in config manager menu. When the Monitor is configured with
NELLCOR hardware, only NELLCOR config manager menu is available; when
the Monitor is configured with MASIMO SET hardware, MASIMO SET config
manager menu is available.
The SpO2 processor monitors the pulse oximetry signal. The processor takes the
signals and derives the oxygen saturation and heart rate data and converts
them into serial data. The serial data from the SpO2 processor is repacked by
DAS board and sent across an isolation barrier (opto couplers) and passed to the
Primary Processor via a dual-channel UART.
UI PWAs
The UI PWAs provide access to the basic functions of the Monitor. The buttons
that control each function are integrated to form a touch pad front panel. The
AC LED is continuously green when an external power source is present or
flashing green when the unit does not recognize the battery. The battery LED is
continuously yellow when the unit is running on battery and flashes yellow
when the battery is low.
DAS PWA
The DAS PWA is composed of ECG and MPDAS. Both ECG and MPDAS are
isolated to earth, and isolated from each other.
ECG accepts signals from a 3-electrode cable or a 5-electrode cable for
processing. The 3-electrode cable provides a single lead configuration with Lead
I, Lead II, or Lead III available. The 5-electrode cable provides a configuration
with Lead I, II, III, aVR, aVL, aVF, VA and displays this waveform as the primary
lead. In 5-electrode configuration, another waveform can be displayed as the
secondary lead.
A hybrid on the PWA provides lead-to-lead defibrillator protection. In addition, a
passive R/C network located on this PWA provides the first stage of high
frequency filtering for EMC and ESU interference rejection. Two electrodes are
selected for ECG measurement by a multiplexer (LS0, LS1 signal controlled) and
passed to a differential amplifier. A second multiplexor selects the third
electrode (the one not sent to the differential amplifier) and drives the signal with
an amplified and inverted version of the common mode voltage of the two
measuring electrodes. This feedback action cancels most of the common mode
signal applied to the differential amplifier. The output signal from the differential
amplifier is then routed to the bandpass filter and pacemaker detection circuit.
The ECG PWA uses the pacemaker detection circuit to prevent pacemaker
signals from interfering with heart rate measurements. The ECG signals are sent
2-12
Revision C
Pneumatic Control
The pneumatic functional block includes the control signal decode logic, the
valve driver circuitry, the pump driver circuitry, and a safety interlock circuit.
There are two transducers on board, PT1 and PT2. PT1 is used for main readings
while PT2 confirms readings and is used to derive overpressure signals. The
following signals are multiplexed into a 16-bit SAR A/D converter via a
multiplexor:
TH REF - the voltage that the amplified PT2 has to attain before the
safety circuit cuts in.
Valve Sense
The 16-bit value out of the ADC is available on the data bus at D[15:0].
Revision C
2-13
LCD Assembly
The Monitor series uses a TFT active-matrix-color liquid display. The 10.4 inch
diagonal color display area contains 640 x 480 pixels and is backlit by coldcathode fluorescent lamps.
The LCD has the following specific characteristics. These are neither defects nor
malfunctions:
The ambient temperature may affect the display condition of the LCD.
The LCD is driven from the Primary Processor via buffers (HCT244) on a
dedicated LCD driver port:
Signal
Name
Clk
Clock
Vsync
Vertical Sync
Hsync
Horizontal
Sync
R[0:3]
G[0:3]
B[0:3]
The display module has a 31-way control signal connector and a 3-way
backlight driving connector.
2-14
Revision C
Printer (Optional)
The Monitor uses a thermal graphics printer. The printer requires a 5 V supply for
its logic circuitry and 12 V (nominal) for the motor. The power and data lines are
connected to the DC2DC Board by a 40 pin connector. The data lines are
connected to the SCC3 port on the Primary Processor.
The printer has a built-in sensor to monitor the printer paper level. When the
printer is out of paper, it sends a PAPER OUT signal to the Secondary Processor.
Revision C
2-15
Interface
Panel
TURBO
TEMP
BOARD
AC MAIN
FAN
SPEAKER
ECG
Analog Output
N I BP
Pneumatic
connector
SpO2
TURBO TEMP
PROBE
OPTICAL
SENSOR
TURBO
HOLDER
Buffer
/MUX
Front
Amp
SOUND
D/A
ECG
MCU
A/D
Isolation Barrier
MPDAS
MCU
ECG/Respiration
Pace
Detector
Amplifier
LP filter
Isolation Barrier
PSU Module
DC2DC Board
Regulated
+5V/+3.3V
Regulated
+12V
Host communication
port
Isolated
RS232
Network Module
NI BP PUMP
AIR HOSE
Power/Data
Non-isolate
Battery
Serial 2
Isolation Barrier
Ethernet
CHARGER /
FUEL GAUGE
PRINTER
(OPTIONAL)
Isolated
MPDAS
Power
Supplies
Isolated
ECG
Power
Supplies
DAS board
MPDAS
Temperature
Respiration
Modulator
Respiration
Demodulator
Driven
GENERATOR
H
y
b
r
i
d
PT1 PT2
UARTS
DC-DC
3.3V REGULATOR
5V REGULATOR
SECONDARY
PROCESSOR
PRIMARY
PROCESSOR
REAL TIME
CLOCK
EEPROM
SRAM
DRAM
FLASH ROM
BOOT ROM
Mainboard
SERIAL
INTERFACE
PORTS
NIBP signal
processing
MANIFOLD
NBP Valve1
3/5
NBP Valve2
2-16
LEADWIRE
ECG
ENCODER
UI_RIGHT
BOARD
CCFL
TFT LCD
UI_LEFT
BOARD
Button
button
Revision C
Revision C
Installation
3-1
3-2
Revision C
Installation: Connections
Connections
Rear Panel Connections
The back of the Monitor has all ports for equipment.
NOTE: When using the remote alarm, the Monitor should be considered the
primary alarm source. The remote alarm is used for secondary
consideration only.
Power Up
After making all connections, plug the power cord into an AC wall outlet.
You can turn on the Dash 2500 Monitor by pressing either the Power button or
the Trim Knob. If the Monitor is using battery power, it can be turned on only by
the Power button on the front of the unit.
On power-up, the Monitor performs the following steps in the order listed:
Revision C
The Monitor performs a self-test - the screen displays the model and
revision number of the Monitor - and a start-up tone sounds.
3-3
Set Up/Configure
Setting up the Dash 2500 Monitor for the First Time
Unpacking and Preparation for Installation
Unpack and identify the contents of all shipping materials:
2.
3.
Unpack the AC cord. Plug the AC cord into the AC Mains input at the back of
the Monitor.
4.
5.
3-4
1.
2.
Use the Trim Knob to select the no option when the Monitor prompts to
admit a new patient.
3.
Revision C
4.
Turn the Trim Knob to scroll down the menu. The arrow at the bottom of the
list indicates that the list continues on a second screen. Highlight the other
system settings option and press the Trim Knob.
5.
Turn the Trim Knob to scroll down the menu to highlight the Adjust date &
time option. Press the Trim Knob to continue.
6.
Turn the Trim Knob to scroll down and highlight the appropriate date and
time components to be changed if necessary (Month, Day, Year, Hour,
Minute, Second). Press the Trim Knob. The field is displayed in a box. Turn
the Trim Knob to the desired number and press the Trim Knob.
7.
After all of the settings are changed, use the Trim Knob to scroll down and
highlight the set new time and date option. Press the Trim Knob to save the
settings and continue.
8.
The message, CAUTION! This will delete all trends, and stored
waveforms. Are you sure you want to do this? displays. Highlight the yes
option and press the Trim Knob. A pop-up window displays the message,
Clearing all trends, and stored waveforms. to confirm that the function is
processing.
Configure
Refer to Appendix C to configure and set up the Dash 2500 Monitor for patient
use.
Mounting Options
Please contact your local dealer and GE Medical systems Information
Technologies for help.
Software Upgrade
Please contact your local GE service provider to upgrade the lastest version of
software.
CAUTION
The monitor indicates a MEMORY LOSS when upgrade the
software. It causes the monitor to revert to factory default
settings. All the user settings, system histories and patients
data will be lost.
Refer to the Dash 2500 Software Download Instruction (PN 2039601-001) for the
procedure of software downloading.
Network set up
Refer to Chpater F Connection for network installation configuration procedure.
Revision C
3-5
3-6
Revision C
4-1
Maintenance
Revision C
4-2
Revision C
Maintenance Schedule
Manufacturer Recommendations
To ensure the Monitor is always functional when required, qualified service
personnel should perform the following regular maintenance.
Cleaning: Clean the unit upon receipt of the equipment, every 24 months
thereafter, and each time the unit is serviced.
Calibrating the NIBP Software: Calibrate the software upon receipt of the
equipment, every 24 months thereafter, and each time the unit is opened
for service.
Electrical Safety Tests: Perform safety tests upon receipt of the equipment,
every 24 month thereafter, and each time the unit is serviced.
Manufacturer Responsibility
CAUTION
Failure on the part of all responsible individuals, hospitals or
institutions, employing the use of this device, to implement the
recommended maintenance schedule may cause equipment
failure. The manufacturer does not, in any manner, assume the
responsibility for performing the recommended maintenance
schedule, unless an Equipment Maintenance Agreement exists.
The sole responsibility rests with the individuals, hospitals, or
institutions utilizing the device.
Visual Inspection
The Monitor and its components should be carefully inspected prior to
installation, once every 24 months thereafter and each time the equipment is
serviced.
Revision C
Carefully inspect the equipment for physical damage to the case, the
display screen, and the keypad. Do not use the Monitor if damage is
determined. Refer damaged equipment to qualified service personnel.
Inspect all external connections for loose connectors or frayed cables. Have
any damaged connectors or cables replaced by qualified service personnel.
4-3
Cleaning
Cleaning the Monitor
Monitor Exterior
The exterior surfaces of the Dash 2500 Monitors may be cleaned with a
dampened, lint-free cloth. Use one of the following approved solutions:
Acetone
Ketone
Betadine
Quaternary ammonium disinfectants such as Virex, Sani-Wipes, AsceptiWipes, or products containing similar active ingredients to these should be
avoided.
Display
To clean the display screen, use a soft, clean cloth dampened with a glass
cleaner. Never spray the glass cleaner directly onto the display, and never use
alcohol or hospital disinfectants like Cidex or Betadine.
Failure to follow these cleaning recommendations may melt, distort, or dull the
finish of displays and cases; blur lettering on labels; embrittle cases and lead to
cracks and breakage; or cause equipment failures. Use of non-approved
cleaning agents is considered abuse and is not covered under warranty.
4-4
Revision C
Materials
Distilled water
Spray bottles
Procedure
1.
2.
Spray the detergent liberally on device. If the material is dried on, allow the
cuff to sit for 1 minute. For soil on the soft part of the closure or the cuff
itself, wipe the material off with a soft cloth. For persistent contamination on
the soft part of the closure, use a soft-bristled brush to loosen particles.
Rinse with copious amounts of distilled water. Repeat until no visible
contamination remains. For soil on the hook part of the closure, use a softbristled brush to remove the material, and rinse with copious amounts of
distilled water. Repeat until no visible contamination remains.
3.
Spray the 10% bleach solution on the affected area until the area is
saturated. Allow the cuff to sit for 5 minutes.
4.
Wipe away any excess solution and rinse the cuff again with distilled water.
Allow 2 hours for drying.
The user has the responsibility to validate any deviations from the
recommended method of cleaning and disinfection.
For additional information on infection control procedures, contact GE Medical
Systems Information Technologies Technical Support.
Temperature Devices
Do not immerse predictive temperature probes. The probe can be cleaned with a
solution of 10% bleach in water. Use a cloth or sponge-just damp, not wet-and
avoid getting any liquid into the interior of the probe.
SpO2 Sensors
Adhesive sensors are sterile and for single use only. Reusable sensors should be
cleaned before reuse with a 70% alcohol solution. If low-level disinfection is
required, use a 1:10 bleach solution. Do not use undiluted bleach (5% - 5.25%
sodium chlorite) or any cleaning solution other than those recommended here
because permanent damage to the sensor could occur. Do not sterilize the
Revision C
4-5
Saturate a clean, dry gauze pad with the cleaning solution. Wipe all surfaces
of the sensor and cable with this gauze pad.
2.
Saturate another clean, dry gauze pad with sterile or distilled water. Wipe all
surfaces of the sensor and cable with this gauze pad.
3.
Dry the sensor and cable by wiping all surfaces with a clean, dry gauze pad.
Periodic Maintenance
Integrity of Hoses and Cuffs
When the pneumatic integrity of any NIBP cuff or hose is in doubt, replace the
cuff and hose, and discard the questionable accessories.
Long-Term Storage
If it becomes necessary to store the Monitor for an extended period of time,
remove all attached accessories. Attach the original packing inserts, and place
the Monitor into the original shipping container.
Generally, long-term storage of a nickel-metal hydride battery in either a
charged or discharged condition has no permanent effect on capacity. Capacity
loss due to self-discharge is reversible, and nickel-metal hydride batteries can
recover to full capacity by proper recharging. For example, cycling through
repeated charge/discharge cycles can restore a full capacity of a nickel-metal
hydride battery that was stored at room temperature for up to one year.
Long-term storage at high temperatures can lead to deterioration of seals and
separators and should be avoided.
Periodic Procedures
Perform the test procedures described in every 24 months, or whenever the
accuracy is in doubt.
NOTE:
An internal, 3.6 V NiMH battery acts as an alarm backup and maintains the
nonvolatile RAM memory when the Monitor is off or away from AC mains. A
system alarm message is generated if backup battery replacement is
required.
4-6
Revision C
Battery Charging
The Monitor charges the NiMH battery whenever the AC power supply is in use.
The Monitor automatically senses if the battery needs recharging. Battery
charging continues whenever it's needed while the Monitor is connected to the
AC power supply, even when the Monitor is turned off.
The battery should be charged at room temperature (59F - 86F; 16C 30C).
It is normal for the battery to become warm during charging or after use.
If the Monitor is idle for extended periods, it should be at least fully charged
and discharged once a month to ensure optimum performance.
Battery Maintenance
Battery should be maintained to prolong its life. Do not monitor patients during
maintenance. Maintenance procedures are as follows:
Revision C
Connect the power cord to the Monitor and the AC wall outlet.
Power the Monitor off and leave attached to AC power for a minimum of 4
hours (8 hours recommended).
Power the Monitor on and verify the battery charge status. If further
charging is required, power the Monitor off and recheck every 30 minutes
(or longer) until the battery is fully charged.
4-7
Maintenance:
Battery Troubleshooting
Trouble
Problem Cause
Remedy
Battery inoperative
or does not last
very long.
Charging battery in
unusually cold or hot
temperatures.
Test Equipment
The recommended test equipment required to perform electrical safety tests is
listed below.
Item
Specification
2006036-001
4-8
Revision C
Revision C
4-9
4-10
Revision C
Revision C
4-11
Patient (Sink) Leakage Current Testing (Mains Voltage on the Applied Part)
This procedure measures the leakage current from a mains voltage source
into the ECG/RESP connector or the SpO2 connector or the TurboTemp
connector of the device. .
4-12
Revision C
Checkout Procedures
These checkout procedures provide service personnel with a method to verify
operational and functional performance of the Monitor. Failure to attain any of
the listed results indicates a potential malfunction of the Monitor.
Perform the checkout procedures when you receive the Monitor, every 24
months thereafter, and each time you service the unit.
The checkout procedures are based on the assumption that the tested monitor
has known good cables and test equipment. It also requires that the user be
familiar with the operation of all test equipment required for the checkout
procedures. For more information concerning the operation of these
components, refer to the operation manual.
NOTE: All devices are tested and calibrated during manufacturing and are
certified for operation at installation.
Revision C
4-13
4-14
Revision C
2.
Use the Trim Knob to select the no option when the Monitor prompts to
admit a new patient.
3.
4.
Turn the Trim Knob to scroll down the menu. The arrow at the bottom of the
list indicates that the list continues on a second screen. Highlight the other
system settings option and press the Trim Knob.
5.
Highlight the go to service mode option and press the Trim Knob. Turn the
Trim Knob and press the Trim Knob again to answer yes at the prompt to
display the dialog box.
6.
A row of numbers is displayed at the bottom of the screen. Turn the Trim
Knob and move the arrow to the desired number, then press the Trim Knob
to select the number. Enter the Service Mode password, 2213.
7.
After the password is selected, turn the Trim Knob to the DONE option and
press the Trim Knob.
8.
In the process of entering the Service Mode, the Monitor resets itself.
Successful entry into the Service Mode is indicated by the Service Menu title
displayed on the upper left side of the display.
NOTE:
The Service Mode can also be entered directly from a cold start by
pressing and holding the following two keys until full power-up: Power
and NIBP Auto. To make any changes to the Service Menu, the
password has to be entered, follow Step 6.
9.
Revision C
At this point the Service Mode main screen should be present in the main
display, as shown in the following graphic. The Service Menu service
parameters area displays a list that corresponds to the number and type of
parameters that have been detected by the Monitor. If the Service Mode
was entered directly (as described in the NOTE above), enter service
password appears above the service parameters on the Service Menu. The
password MUST be entered (as described in Steps 5 and 6) before any
changes to calibration can be made.
4-15
SpO2 Testing
For Monitors With NELLCOR SpO2 NELLCOR recommends use of the SRCMAX Portable Tester for use with the Dash 2500 Monitor equipped with the
NELLCOR SpO2 system.
On occasion when testing the integrity of the NELLCOR oximetry system,
abnormal results may occur when introducing large changes in the pulse rate
and/or pulse amplitude. Extreme changes in the rate sent to the NELLCOR
sensor by the SpO2 simulator may cause the SpO2 algorithm to completely miss
finding the pulse rate. This is an expected result. To work around this,
incrementally step up or down the settings on your SpO2 simulator and allow
the Monitor to detect and display the new pulse rate or saturation.
For Monitors With MASIMO SET SpO2MASIMO SET recommends BIO-TEK
SpO2 simulators.
4-16
Revision C
Test Procedure
The following table shows the allowable tolerance of the indicated simulator
values.
Range
Accuracy
2 digits
3 digits
1% - 69%
No accuracy
required
The following procedure applies to both Nellcor and Masimo equipped units.
1.
Disconnect all sensor cables from the SpO2 Parameter, and ensure that the
SpO2 parameter is listed within the main Service Menu.
2.
From the Service Menu, turn and press the Trim Knob to select the SpO2
service parameter. The SpO2 service menu appears. The text under Error
and Version sections reflects the installed type of SpO2. The illustration
shows both text examples.
Revision C
3.
All SpO2 mode operations take place with MASIMO SET and NELLCOR
power-up defaults. No menu settings are reflected.
4.
Connect the appropriate SpO2 simulator and cable to the SpO2 parameter.
Be sure it is fully seated in the socket.
4-17
Vary the values on the simulator. Verify that the Monitor responds
accordingly by displaying the proper heart rate value and saturation value.
NIBP Testing
4-18
Revision C
Using the calibration kit, an adult cuff and air hose, and a manometer, set up
the equipment as shown in the NIBP Test Setup graphic. Connect the hose
to the NIBP Parameter. Make sure that all of the fittings are tight and that
the valve on the manual inflation bulb is fully closed.
Revision C
2.
From the Service Menu, turn and press the Trim Knob to select the NIBP
service parameter.
3.
Turn and press the Trim Knob to select start leak test. Observe that the
Leak Test Status message on the menu indicates Busy.
4.
Observe that the pump begins inflating the system to 200 - 210 mmHg, at
which point the pump operation will cease. The Monitor will begin to
calculate system pressure loss rate.
5.
After about 60 seconds the pressure is released, and the menu should
display Leak Test Status Passed, and the Leak Test Results indication
should be a value less than 6. Service Error: None should continue to
display.
6.
If the menu displays Leak Test Failed, continue with Steps 7 through 9.
4-19
Using the calibration kit, an adult cuff and air hose, and a manometer, set up
the equipment as shown in the Leak Test Setup graphic.
Close the pressure release valve on the manometer inflation bulb and slowly
increase the pressure to 200 mmHg 1 mmHg.
9.
4-20
1.
Using the calibration kit, an adult cuff and air hose, and a manometer, set up
the equipment as shown in the NIBP Test Setup graphic. Connect the hose
to the NIBP Parameter. Make sure all fittings are tight and that the inflation
bulb valve is closed tightly.
2.
From the Service Menu, turn and press the Trim Knob to select the NIBP
service parameter.
3.
4.
5.
Observe that both PT1 Pressure and PT2 Pressure equal initial values of 0
mmHg.
6.
7.
Fold the adult cuff so the index line is aligned with the inner range mark on
the inside of the cuff. Make sure all fittings are tight and that the valve on the
inflation bulb is closed tightly. If there is doubt about the integrity of the
system, perform the leak test before continuing.
8.
Close the pressure release valve on the manometer inflation bulb and
manually pump up the pressure until the manometer indicates
approximately 220 mmHg.
Revision C
Allow the pressure to stabilize for at least 1 minute. Then open the pressure
release valve on the manometer inflation bulb and carefully bleed off
pressure until the manometer indicates 200 mmHg.
10. Observe that the values of PT1 Pressure and PT2 Pressure on the menu
indicate within 1 mmHg of the pressure shown on the manometer. If not,
please check and repair the device.
11. Verify the system linearity by repeating Steps 8 and 9 using manometer
readings of 250 mmHg, 150 mmHg, and 50 mmHg. Observe that the PT1
and PT2 Pressures are within 3 mmHg of manometer readings for each of
these pressure indications. If not, proceed to the Pressure Recalibration
section.
Pressure Recalibration
1.
Using the calibration kit, an adult cuff and air hose, and a manometer, set up
the equipment as shown in the NIBP Test Setup graphic. Do not connect the
pneumatic hose to the NIBP port yet.
2.
From the Service Menu, Turn and press the Trim Knob to select the NIBP
service parameter.
3.
4.
5.
Observe that both PT1 Pressure and PT2 Pressure display initial values of 0
on the menu.
6.
Turn and press the Trim Knob to select cal press zero. Observe that the
message Inflate System to 200 mmHg Then Hit Cal Press 200 is displayed
on menu.
7.
8.
Fold the adult cuff so the index line is aligned with the inner range mark on
the inside of the cuff. Make sure all fittings are tight and that the valve on the
inflation bulb is closed tightly. If there is doubt about the integrity of the
system, perform the leak test before continuing.
9.
Close the pressure release valve on the manometer inflation bulb and
manually pump up the pressure until the manometer indicates
approximately 220 mmHg.
10. Allow the pressure to stabilize for at least 1 minute. Then open the pressure
release valve on the manometer inflation bulb and carefully bleed off
pressure until the manometer indicates a little more than 200 mmHg.
11. Turn and press the Trim Knob to select cal press 200, but do not press the
Trim Knob.
12. When the manometer indicates exactly 200 mmHg, press the Trim Knob.
Observe that system pressure is released, and the message, !!!!! CAL INFO
NOT SAVED!!!!! Service Error: None is displayed on menu.
13. Turn and press the Trim Knob to select save cal info. The message, !!!!!CAL
INFO NOT SAVED!!!! disappears. This indicates your new calibration values
have been saved.
14. Repeat the calibration check procedure to confirm the calibration setting.
Revision C
4-21
Overpressure Tests
1.
Using the calibration kit, an adult cuff and air hose, and a manometer, set up
the equipment as shown in the NIBP Test Setup graphic. Connect the hose
to the NIBP Parameter. Make sure all fittings are tight and that valve on
inflation bulb is closed tightly.
2.
From the Service Menu, Turn and press the Trim Knob to select the NIBP
service parameter.
3.
4.
5.
6.
Turn and press the Trim Knob to select inflate on. The pump should begin to
inflate the system.
7.
8.
9.
10. Turn and press the Trim Knob to select neo ovp select. Verify the menu
displays Overpressure Latch: Cleared and Overpressure Selected: Neo.
11. Turn and press the Trim Knob to select inflate on. The pump should begin to
inflate the system.
12. Watch the pressure indication increase on the manometer, and observe
that the pump is shut down and the pressure is released when the
manometer indicates in the range of 150 to 165 mmHg. Observe that the
menu displays Service Error: Over Pressure Condition Detected.
13. If the overpressure test results in an out of tolerance condition, contact GE
Medical Systems Information Technologies Technical Support at 1-800-5587044 (USA), 86-800-810-8188 (China) or contact your local representative. If
you are a cell phone or Xiaolingtong user, or your area does not support 800
free-call service, please dial: 86-010-67882652 (China). If you are a EMEA
user, please contact your local GE service provider.
Overpressure Calibration
NOTE:
Overpressure calibration is adjusted by software based on Pressure
Calibration section.
4-22
Revision C
ECG Testing
Connect the ECG leads to the ECG trunk cable prior to connection to the Monitor.
The simplest way to function test the ECG circuitry is through the usage of an
ECG simulator with the Monitor in normal monitoring mode.
1.
2.
3.
Select no at new patient prompt. Set the ECG high alarm to 150 and the low
alarm to 50.
4.
5.
From ECG menu, select Pace 1 and verify paced marker on display
waveform.
6.
From ECG menu, select Pace 2 and verify paced marker on display
waveform.
7.
From ECG menu, select PACE OFF. Turn paced off on simulator.
8.
After unit has learned the patient waveform change the bpm to 30.
9.
4-23
RESP Testing
1.
2.
3.
4.
5.
6.
7.
8.
4-24
Revision C
1.
Turn Monitor off. Make sure the temp probe is properly stored in the probe
well.
2.
3.
4.
5.
partially remove the probe from the well (stop when you hear the
Monitor beep).
b.
Quickly re-insert the probe and remove again (you should hear two
beeps of a different tone).
c.
6.
Record and verify the reading in the temp display is 80.2 0.2F.
7.
8.
Record and verify the reading in the temp display is 98.6 0.2F.
9.
10. Record and verify the reading in the temp display is 107.8 0.2F.
11. Calibration is complete. If the monitor does no pass the calibration
verification then contact Technical Support.
Revision C
1.
Ensure that paper has been loaded into the Recorder and you are presently
in the Service Mode.
2.
From the Service Menu, turn and press the Trim Knob to select the Recorder
test option. Turn and press the Trim Knob to select the Print 1 Waveforms
4-25
Allow for the paper to spool out a 12-inch printed section then press Stop
Test.
4.
Select Vertical Text test. Verify that the printed text is legible and evenly
spaced.
Select Horizontal Text test. Verify that the printed text is legible and evenly
spaced.
4-26
Revision C
Battery Testing
From within the Service Menu, battery status information is displayed on the
upper right-hand section of the display.
Battery Health
The Monitor's software approximates the true status of the battery's health. The
value indicated is displayed as both a number (in percentage) on the top of the
display and as an icon on the lower right area of the display.
Charge Type
FAST indicates battery is charged fast when it is not full. TRICLE indicates
battery is charged slowly when it is already full.
Battery Failed
Any result other than FALSE indicates that the internal battery has suffered a
failure and should be investigated.
Revision C
4-27
Test Procedure
1.
Verify AC Mains indicator on front panel of unit near Power hardkey is lit
with AC Mains plugged in.
2.
Turn on Monitor.
3.
4.
Turn the Trim Knob to the yes option and press the Trim Knob.
2.
After 2 seconds, the system freezes, an alarm sounds, and the screen goes
blank. Recycle the system power using the Power hardkey. To return to the
Service Mode, repeat the procedures as described in Service Mode section.
1.
From the Service Menu, turn and press the Trim Knob to select keypad KEY
test. With the exception of the Power hardkey, verify that each key press
produces a tone. You may hear different tones in some occasions. This does
not affect the testing results.
2.
After all keys have been tested, press the Trim Knob again to stop the test.
1.
From the Service Menu, turn and press the Trim Knob to select Sound Test.
2.
Select ON to start the test. The monitor generates a serial of tones the
Monitor has one by one every 2 seconds.
3.
Sound Testing
4-28
Revision C
Communications Testing
Set up Terminal
1.
2.
Turn the Trim Knob to get to the Main Menu and select other system
settings.
2.
3.
4.
After the unit reboots, turn the Trim Knob to display the Configuration
Menu.
5.
6.
7.
Select Serial 2 setup and configure Serial 2 for ASCII cmd, 9600 baud.
8.
9.
Communication Test
Execute the following commands (by sending text files from the terminal
program) and verify the appropriate response.
NOTE:
Each string is preceded by a space. "^" represents the space character.
"^NC0!E" Verify that UUT pump starts.
"^ND!5" Verify that UUT pump stops.
"^TB!9" Verify return temperature status in the form "...TB-99999...".
Revision C
4-29
Install the test plug into 15 pin communications port on rear of unit, as
shown in Test Plug Assembly graphic.
2.
Use the DMM to measure voltage between pins 4 and 8 of DB15 connector
and record the result.
REMOTE ALARM TEST PLUG
15
1
14
2
13
3
12
4
11
5
10
6
9
7
470
15 PIN MALE (FRONT VIEW)
From the Service Menu, turn and press the Trim Knob to select Alarm relay.
4.
Select Alarm relay / ON. Measure and record voltage between pins 4 and 8
of DB15 connector. Verify the voltage is 0 V.
5.
Select Alarm relay / OFF. Measure and record voltage between pins 4 and 8
of DB15 connector. Verify the voltage is 5 V.
6.
Network Testing
Install and set up the network module for Dash 2500. Check that the connector
and the plugs are clean and intact, then connect the monitor into the network.
Select View patient tab on the CIC, set up the unit number and bed number of
the connected monitor to establish communication. If the unit name is the same
as CIC unit setting on the monitor, the CIC will establish communication
automatically. If it does not display automatically, right click an multi-patient
view window and select the correct unit/bed of the monitor.
NOTE: In general, it needs less than 30 seconds to let monitor to be detected
and establish communication by CIC after you finished all above steps.
If the monitor information (e.g. bed number, unit number, and patient name) is
displayed on the CIC patient tab, the communication is established successfully.
Country of Use
This option have two setting values: CHINA and Other. The CHINA setting will
cause the monitor to have the same factory default, ECG parameter, QRS width
setting between adult, pediatric, and neonate patient types. The user can still
choose to change the setting for a particular patient through the ECG parameter
advanced settings menu.
4-30
Revision C
Revision C
4-31
4-32
Revision C
Revision C
Troubleshooting
5-1
5-2
Revision C
Revision C
Adult mandrel, end block and spacer blocks (DNI PN 5215-0268, 5215-0269)
(Lead test)
5-3
General Troubleshooting
Problem
Solutions
No AC power
Check AC power
Faulty cables
Faulty display
Faulty encoder
5-4
Cause
Revision C
Parameter Troubleshooting
Problem
Revision C
Cause
Solutions
Faulty pneumatics
assembly
Faulty printer
5-5
Error Codes
Alarm Code Interpretation
If any other alarms appear that are not listed in the paragraphs that follow,
record the error message and report the failure to Customer Support. Refer to
the operation manual for information about patient alarms and general
procedural alarms.
For network modules troubleshooting, the information appear as the LED light in
the network module board, refer to Appendix F, LEDs .
System Failures
When a system failure is encountered, the error code is displayed on the screen
for 5 seconds and the system enters fail-safe mode. The error code is recorded
in the history log.
General system error codes are listed below. If any other SY or similar code
displays, report it to Customer Support.
5-6
Explanation
SY-16
SY-19
SY-20
SY-40
Unexpected interrupt
SY-43
SY-44
Revision C
Explanation
8193
8202
8222
8232
8252
8253
26631
27268
Revision C
Explanation
110
130
131
140
141
142
150
180
Excessive leakage
190
220
221
227
5-7
Explanation
101
MPDAS Board
102
MPDAS Board
103
MPDAS Board
109
113
MPDAS Board
114
MPDAS Board
128
MPDAS Board
128-132
MPDAS Board
MPDAS Board
201
5-8
Explanation
101
MPDAS Board
102
MPDAS Board
103
MPDAS Board
111
MPDAS Board
120
MPDAS Board
Revision C
Revision C
Explanation
Suggested Replacement
101
MPDAS Board
102
MPDAS Board
103
MPDAS Board
125
MPDAS Board,
Main CPU Board
126
MPDAS Board
129
MPDAS Board
130
MPDAS Board
131
MPDAS Board
132
MPDAS Board
133
MPDAS Board
134
MPDAS Board
135
MPDAS Board
136
MPDAS Board
137
MPDAS Board
138
MPDAS Board
139
MPDAS Board
140
MPDAS Board
141
MPDAS Board
150
MPDAS Board
151
MPDAS Board
5-9
Explanation
101
MPDAS Board
MPDAS Board
102
103
104
5-10
Explanation
Suggested Replacement
101
Printer
102
Printer
103
Printer
104
Printer
105
Printer
110
Printer
111
Printer
112
Printer
114
Bad command
Printer
115
Bad command
Printer
120
Printer
121
Printer
122
Invalid location
Printer
123
Printer
140
Printer
141
Printer
Revision C
Revision C
Field Replaceable
Units
6-1
6-2
Revision C
Ordering Parts
The parts lists and drawings in this chapter supply enough detail for you to order
parts for the assemblies considered field replaceable.
If you require additional information or troubleshooting assistance, contact GE
Technical Support.
To order parts, contact Service Parts at the address or telephone number listed
on the How to Reach Us... page found in service manual CD.
For the latest parts information, including substitutions, obsolescence, and
compatibility, please contact you local GE service provider.
Revision C
6-3
Parts
The table below lists field replaceable units that can be ordered.
NOTE:
Due to recent branding changes, the part ordered may vary slightly from
the previous part. Contact you local GE service provider for a full list of
available parts
Find Number
6-4
Item Number
Item Description
633176
2023852-002
2023852-049
2023852-050
2023852-005
2023852-006
2023852-007
2023852-008
2023852-009
10
2023852-010
11
2023852-011
12
2023852-012
13
2023852-013
14
2023852-014
15
2023852-015
16
2023852-016
17
2023852-051
18
2023852-018
19
2023852-019
20
2023852-020
21
2023852-021
22
2023852-022
23
2023852-023
Revision C
Find Number
Revision C
Item Number
Item Description
24
2023852-025
25
2023852-026
26
2023852-027
27
2023852-028
28
2023852-029
29
2023852-030
30
2023852-031
31
2023852-032
32
2023852-037
33
2023852-039
34
2023852-040
35
2023852-041
36
2023852-042
37
2023852-043
38
2023852-044
39
2023852-045
40
2023852-046
41
2023852-047
42
2023852-048
43
2023852-052
44
2023852-054
45
2023852-056
6-5
Assembly Drawings
The assembly drawings below indicate the field replaceable parts.
6-6
Revision C
Revision C
6-7
6-8
Revision C
Revision C
6-9
6-10
Revision C
Revision C
6-11
6-12
Revision C
Revision C
6-13
Disassembly Guidelines
REPAIR TO THE FRU LEVEL Field repairs are recommended to the field
replaceable unit (FRU) only. Attempting a field repair on a PCB or a factory
sealed component or assembly could jeopardize the safe and effective
operation of the Monitor.
DAS ASSEMBLY Do NOT open the DAS assembly as this breaks the isolation
barrier which may result in patient death or serious injury. The DAS assembly is
a field replaceable unit only. There are NO field repairs or adjustments for the
DAS assembly.
DC2DC ASSEMBLY When removing the GCX plate from the bottom of the
Monitor, clearly identify the screws to ensure the same screws are used to
replace the GCX plate. Screws that are too long will penetrate into the DC2DC
assembly and cause the battery to split or break.
NOTE:
GE recommends that you assemble the Monitor using the NEW fasteners
(screws, washers, etc.) provided in the FRU Kits. Some fasteners, like the
screws with a thread locking coating, are NOT intended to be re-used more
than three times.
Tools Required
A standard set of hand tools is required for disassembly and assembly.
Before Disassembly
Before you disassemble the Monitor, you should ALWAYS do the following tasks.
6-14
1.
2.
3.
4.
Revision C
Hardware Precautions
When disassembling the Monitor, observe the following guidelines:
Remove the handle assembly, then remove the display assembly to access
the field replaceable units of the display assembly and the main unit.
Note the positions of wires, cables, and different sized screws; marking them
if necessary to ensure they are replaced correctly.
Discharge any static charge you may have built up before handling
semiconductors or assemblies containing semiconductors.
These guidelines may not guarantee a 100% static-free workstation, but can
greatly reduce the potential for failure of any electrical/electronic assemblies
being serviced.
Revision C
6-15
Remove/Replace Battery
6-16
1.
Remove the four M5 screws with the driver, then take off the aluminium
plate.
2.
3.
Release the lock and then take out the battery and wires.
4.
Revision C
Revision C
1.
2.
Remove all the six screws from rear case. The unit is separated into front
case and rear case.
3.
Take off the four different wires and replace the front case.
6-17
Separate the front case and the rear case, then remove the four screws and
take the Main Board out of rear case.
2.
Pull out all tubes that are connected to the Main Board and replace the
Main Board. Connect the jumper to the Main Board as indicated in the
following graphic.
3.
Connect all the tubes to the Main Board as indicated in the following
graphics.
Tube Connection
6-18
Revision C
2.
Remove the four screws from the rear case assembly and take out the
chassis assembly.
3.
Pull out the DAS assembly and replace with the new DAS assembly. There
are two DAS assemblies to select: DAS assembly with NELLCOR SpO2 or
DAS assembly with MASIMO SpO2.
Revision C
6-19
6-20
1.
2.
3.
Revision C
Revision C
1.
2.
3.
6-21
6-22
1.
2.
Remove all the six screws and take out the LCD assembly.
3.
4.
Take out the gasket, then use a sleeve to get the corder out. Replace it with
the new corder and replace the knob.
Revision C
Revision C
1.
Pull out the cables and remove all the four screws.
2.
6-23
6-24
1.
2.
3.
Revision C
Pull out the printer holder and slip out the foam gasket.
2.
Replace speaker.
3.
Revision C
1.
2.
6-25
2.
3.
6-26
1.
2.
3.
Revision C
Revision C
1.
2.
6-27
6-28
1.
2.
If the monitor is the old version, replace the ports two screws into new
screw-locks, with the 5mm hexagonal socket wrench.
3.
Revision C
5.
After Assembly
Perform the Electrical Safety Tests and Checkout Procedures after assembly to
ensure the Monitor with new FRUs work functionally. Refer to Chapter 4 and
Appendix B for details.
Revision C
6-29
6-30
Revision C
Revision C
Technical
Specifications
A-1
A-2
Revision C
Technical Specifications
Due to continual product innovation, specifications are subject to change
without notice. The following specifications are accurate as of the date of this
publication, and pertain to the Monitor
General Specifications
Specifications
Mechanical
Monitor
Weight
Dash 2500 Patient Monitor
12 lb (5.5 kg)
Environmental*
Operating Temperature
Operating Humidity
5% to 95% noncondensing
5% to 95% noncondensing
Operating Atmospheric
Pressure
Electrical
AC Input Voltage
100-240 V
AC Input Frequency
50/60 Hz
AC Input Power
120 VA
Internal Battery
Power Supply
The Dash 2500 Patient Monitor can be powered from the internal battery or AC
power.
Battery
An internal, rechargeable battery pack powers the Monitor for greater than 180
minutes at a specified load. The battery typically charges to full capacity within 4
hours when power off.
* The Monitor may not meet Performance Specifications (ANSI/AAMI SP10) if it is stored or used outside of the specified ranges of environmental
parameters.
Revision C
A-3
Specifications
Power Cable
Fuses
Internal
FS1
FS2
FS3
FS4
FS5
External
FS
FS
A-4
Revision C
NIBP Specifications
Specifications
Method
Modes
BP Measurement Ranges
Systolic
MAP
Diastolic
Resolution
1 mmHg
Accuracy
120 s (adult/pediatric)
85 s (neonate)
Pulse Rate
When NIBP is the source, HR values are derived from the pulserate
that is determined by the oscillometric technique of measuring
blood pressure. The rate source field is labeled NIBP.
Adult/Pediatric Range: 30 to 200 bpm ( 3.5% or 3 bpm)
Neonate Range: 30 to 220 bpm ( 3.5% or 3 bpm)
GE Healthcare Patents
4,638,810; 5,052,397; 4,754,761; 5,170,795; 6,188,407; 5,357,970; 5,704,362;
5,680,870; 5,518,000; 6,893,403; 6,423,010; 6,358,213; 5,704,362; 5,579,776 and
international equivalents. US patents pending.
Revision C
A-5
1 to 100%
Pulse Rate
20 to 250 bpm
Perfusion Range
0.03 to 20%
Accuracy
Saturation
Adult*
70 to 100% 2 digits
Neonate*
70 to 100% 3 digits
Low Perfusion**
70 to 100% 2 digits
Pulse Rate
Adult and Neonate
Low Perfusion**
*Adult specifications are shown for OxiMaxMAX-A and MAX-N sensors with the N-600. Saturation accuracy will vary by sensor type.
**Applicability: OxiMax MAX-A, MAX-AL, MAX-P, MAX-I, and MAX-N sensors.
Sensor Model
OXIMAX
MAX-A, MAX-AL
2 digits
MAX-N (Adult)
2 digits
MAX-N (Neonate)
3 digits
MAX-P
2 digits
MAX-I
2 digits
MAX-FAST
2 digits
SC-A (Adult)
2 digits
SC-PR (Neonate)
3 digits
A-6
Revision C
MAX-R
OxiCliq
OxiCliq A
2.5 digits
OxiCliq P
2.5 digits
OxiCliq N (Adult)
2.5 digits
OxiCliq N (Neonate)
3.5 digits
OxiCliq I
2.5 digits
3 digits
D-YS (Neonate)
4 digits
3.5 digits
3.5 digits
DS-100A
3 digits
OXI-A/N (Adult)
3 digits
OXI-A/N (Neonate)
4 digits
OXI-P/I
3 digits
Power Dissipation
The MAX-N, D-YS, OXI-A/N, and OxiCliq N were tested on patients >40 kg.
Revision C
A-7
1 to 100%
Pulse Rate
25 to 240 bpm
Perfusion Range
0.02 to 20%
70 to 100% 2 digits
70 to 100% 3 digits
70 to 100% 3 digits
Low Perfusion
70 to 100% 2 digits
0 to 69% unspecified
Pulse Rate
Without Motion
With Motion
* The MASIMO SET SpO2 parameter with LNOP-Adt sensors has been validated for no motion accuracy in human blood studies on healthy adult
volunteers in induced hypoxia studies in the range of 70-100% SpO2 against a laboratory co-oximeter and ECG monitor. This variation equals plus
or minus one standard deviation. Plus or minus one standard deviation encompasses 68% of the population.
**The MASIMO SET SpO2 parameter with LNOP-Adt sensors has been validated for motion accuracy in human blood studies on healthy adult
volunteers in induced hypoxia studies while performing rubbing and tapping motions at 2 to 4 Hz at an amplitude of 1 to 2 cm and a nonrepetitive
motion between 1 to 5 Hz at an amplitude of 2 to 3 cm in induced hypoxia studies in the range of 70-100% SpO2 against a laboratory co-oximeter
and ECG monitor. This variation equals plus or minus one standard deviation. Plus or minus one standard deviation encompasses 68% of the
population.
The MASIMO SET SpO2 parameter with LNOP-Neo Pt sensors has been validated for neonatal motion accuracy in human blood studies on
neonates while moving the neonates foot at 2 to 4 cm against a laboratory co-oximeter and ECG monitor. This variation equals plus or minus, one
standard deviation. Plus or minus one standard deviation encompasses 68% of the population.
The MASIMO SET SpO2 parameter has been validated for low perfusion accuracy in bench top testing against a Biotek Index 2 simulator and
MASIMOs simulator with signal strengths of greater than 0.02% and a % transmission of greater than 5% for saturations ranging from 70 to
100%. This variation equals plus or minus, one standard deviation. Plus or minus one standard deviation encompasses 68% of the population.
NOTE: Note: LNCS sensors using LNCS cable technology offer nearly the same performance as the LNOP flexcircuit design. The LNOP single patient sensor line is still the industry "gold standard", while LNCS is the best
cabled sensor alternative.
A-8
Revision C
Sensor Model
LNOP
LNOP-ADT
2 digits
LNOP-ADT Long
2 digits
LNOP-PDT
2 digits
LNOP-NEO
3 digits
LNOP-NEO PT
3 digits
LNOP-DCI (reusable)
2 digits
LNOP-DCSC (reusable)
2 digits
LNOP-DCIP (reusable)
2 digits
NRI25 (reusable)
2 digits
Resolution
Saturation (% SpO2)
1%
Interfering Substances
Power Dissipation
MASIMO Patents
5,482,036; 5,490,505; 5,632,272; 5,685,299; 5,758,644; 5,769,785; 6,002,952;
6,036,642; 6,067,462; 6,157,850; 6,206,830; 6,263,222 and international
equivalents.
Revision C
A-9
ECG Specifications
Specifications
Leads available
0.5 to 5.0 mV
30 to 300 bpm
3 bpm or 3% of reading, whichever is greater
1 bpm
Bandwidth
0.5 to 40 Hz +1/-6 dB
0.05 to 40 Hz +1/-6 dB
0.05 to 100 Hz +1/-6 dB
Standardizing voltage
1 mV marker
Input Impedance
Common mode
Differential
> 2.5 M at 10 Hz
> 2.5 M from DC to 60 Hz
60 Hz tolerance
up to 10 mV
Pacemaker detection/rejection
Input voltage range
2 mV to 700 mV
Time to alarm
ANSI/AAMI EC13-2002 Section 4.1.2.1 d). The Heart Rate Averaging computation is as follows, the average of the
last 8 R-to-R intervals, the update rate of the Heart Rate on the display is once per second.
When tested in accordance with ANSI/AAMI EC13-2002 Section 4.1.2.1f), the response time of the heart rate
meter to changes in heart rate is:
Step increase from 80 to 120 BPM
Step decrease from 80 to 40 BPM
When tested in accordance with ANSI/AAMI EC13-2002 Section 4.1.2.1g), the time to alarm for ventricular
tachycardia is:
For Figure 4a
A-10
Revision C
Specifications
For Figure 4b
When tested in accordance with ANSI/AAMI EC13-2002 Section 4.1.4.3), the minimum input slew rate that will
cause approximately 50% of the pulses to trigger the pacer pulse detector is:
For Figure 5d
When tested in accordance with ANSI/AAMI EC13-2002 Section 4.1.2.1e), the heart rate after a 20 second
stabilization period is:
For Figure 3a
For Figure 3b
For Figure 3c
For Figure 3d
RESP Specifications
Specifications
ECG-Derived Respiration Rate
Leads available
I or II
Range
Accuracy
2 breaths/min or 3%
of reading; whichever is greater
Resolution
1 breath/min
Base Impedance
100 to 2000
Detection Sensitivity
Excitation Frequency
65.5 kHz
Amplitude
Revision C
A-11
HR/Pulse Specifications
Specifications
ECG
Heart rate accuracy
30 to 300 bpm
3 bpm or 3% of reading, whichever is greater
Time to alarm
NELLCOR SpO2
Range
20 - 250 bpm
Low Perfusion
MASIMO SpO2
Range
25 - 240 bpm
With Motion
30 - 200 bpm
Neonate
30 - 220 bpm
Accuracy
3.5%
Alarm Limits
Range
A-12
30 - 250 bpm
Revision C
Fahrenheit (F)
Celsius (C)
Predictive Mode
Range
Resolution
0.1F (0.1C)
Monitor Mode
Range
Accuracy
Resolution
0.1F (0.1C)
Probes
CAUTION:
Use only Alaris probes and probe covers. The size, shape, and
thermal characteristics of the probe covers can affect the
performance of the instrument. Inaccurate readings or retention
problems may occur unless Alaris probes and probe covers are
used.
Determination Time
Alaris Patents
D,300,728; D,300,909 and other pending patent.
Revision C
A-13
A-14
Revision C
Revision C
Test Record
B-1
B-2
Revision C
Test Record
Model ____________
TEST RECORD
Serial ____________
Description
Min
Max
Actual
Pass
Fail
N/A
Safety Testing
AC mains ground pin to earth ground resistance ()
0.1
500
500
500
500
500
500
10
50
10
50
100
500
100
500
100
500
100
500
50
50
96
100
76
84
SpO2 Testing
Revision C
B-3
Test Record
Description
Min
Max
Actual
Pass
Fail
N/A
47
53
147
153
247
253
300
330
150
165
161
195
107
133
67
93
76
84
94
151
87
123
52
78
26
34
156
164
176
184
17
23
57
63
B-4
Revision C
Test Record
Description
Min
Max
Actual
Pass
Fail
N/A
Temperature Testing (Perform in Service or Monitor Mode - requires Alaris Temp Simulator)
Measured Temp in F (98.6 nominal)
98.4
98.8
79.9
80.5
107.5
108.1
4.7
5.3
0.0
0.1
Probe Disconnected
Probe Type
Probe In/Out Detect
Recorder Testing
Recorder Test
Battery System Testing (Perform in Monitor Mode)
Verify AC Mains Indicator
Remove AC, Verify uninterrupted battery operation
Verify Battery LED is lit
Fail-safe Logic Testing
Fail-safe Logic
Keypad Key Testing (Perform in Service Mode)
Verify appropriate responses to key presses
Sound Testing
Speaker Test
Communications Testing
Verify pump starts, stops and temp status returns
Remote Alarm Testing
Network Testing
Verify network communication
Tested By:
Date:
Signature:
Facility:
Revision C
B-5
Test Record
B-6
Revision C
Revision C
Configuration
C-1
C-2
Revision C
2.
Select go to config mode. The message, This will initiate the sequence for
entering Configuration Mode. Do you want to do this? appears.
3.
4.
The message, Please enter the Config Mode password. appears. Enter the
password. FACTORY SET CONFIG PASSWORD: 2508.
5.
Select Done.
6.
The system will restart in configuration mode. Press the Trim Knob to access
the Configuration Menu.
Revision C
1.
2.
3.
A popup window appears: All unsaved changes to the current default will
be lost! Are you sure you want to do this? Select Yes.
4.
5.
To save your changes for the selected table, go to other system settings,
select save default changes. The default changes are saved.
C-3
2.
3.
A popup window appears: This will exit configuration mode. All unsaved
changes will be lost. Are you sure you want to do this? Select Yes.
4.
C-4
Revision C
Revision C
User can choose whether to view trended derived vital signs data in the Full
trends window either numerically or graphically. When the Full trends window
is opened, the configured default becomes the confirmed selection of this
choice.
Display as
blank
0 chosen
This opens a data entry dialog with the text Enter Unit Number.
blank
ADULT
N/A
This opens a data entry dialog box with the text Enter Bed Number.
Confirmed selections are printed on the bedside recorder when print chosen
graphs is chosen. If there are no confirmed selections for this choice when
print chosen graphs is confirmed, up to two selections are automatically
confirmed. The selections automatically confirmed are the highest priority
parameters included in the confirmed patients parameter list, up to a
maximum of two.
discharge
It allows the user to adjust the length of time alarms are temporarily silenced
when the alarm silence feature is activated.
Admit Patient
N/A
N/A
It allows the user to autoset all currently operating parameter alarm limits
around the patient's current condition using the percentage confirmed in
Choose autoset %
It opens the Config Adjust Alarms menu.
20%
It allows the user to change the percentage at which limits are automatically
adjusted around the patient's current condition when the autoset all is
confirmed.
It allows the user to adjust the alarm volume. Alarm volume applies to all
alarms (excluding system failures).
Factory Default
It allows the user to specify the lowest range for which the alarm volume can
be adjusted.
autoset all
Choose autoset %
Config settings
Adjust Alarms
Adult
Pediatric
Neonate
C-5
C-6
N/A
yes
numbers
1 min/50 min
It allows the user to turn on or off the viewing of the Mini trends window in the
waveform region of the main monitoring screen.
User can choose whether to view trended derived vital signs data in the Mini
trends window either numerically or graphically.
View vitals every is displayed when Display as in Mini trends is numbers. It
allows the user to choose the set of viewable times for the Mini trends window.
View vitals for is displayed when Display as in Mini trends is graphs. It allows
the user to choose the span of time to be viewed in the Mini trends window.
Display as
User specifies which window (either Full trends or Mini trends) that the hardkey
activates/deactivates.
Advanced settings...
lo
Adjust limits
Allows the user to set the "hi" alarm limit.
User specifies the volume for the QRS beep relative to the other volume
settings for alarms and key clicks.
hi
Select source
Setup HR/Pulse
yes
green
yes
warning
N/A
A=50, P=60
A=150, P=160
auto-set
auto
mini trends
N/A
Config Settings
Mini trends...
It opens the Mini trends menu which has go to main menu as its first menu
choice.
Factory Default
1 min/40 min
Adult
Pediatric
Neonate
Revision C
Revision C
PACE OFF
yes
N/A
It allows the user to instruct the Monitor to analyze ECG data for pacemaker
pulse
User specifies whether the monitor will detect and display arrhythmia
conditions
It allows the user to instruct the Monitor to "learn" a new ECG pattern so that it
can more accurately calculate the patients heart rate
Pacer detection?
Arrhythmia detection
Re-learn
N/A
0.5 to 40 Hz
N/A
crisis
procedural
procedural
User specifies the default sweep speed for all cardiac-based waveforms
(except the ECG waveform if Fixed ECG sweep speed? is yes). This choice
appears in all menus associated with cardiac-based waveforms. A change in
one menu affects all cardiac-based waveforms.
The user chooses this to interject one cycle of a 1 mV square-wave into the
display of the ECG waveform
User specifies the type of display filtering done on raw ECG waveform data
before it is displayed or recorded.
User specifies the size of QRS width detection by the EKPro algorithm.
User specifies audio alarm associated with these alarms.
User specifies the audio alarm associated with ECG VTACH in Adult and
Pediatric modes. Menu item is not viewable in Neonate mode.
User specifies the audio alarm associated with ECG LEAD FAIL.
User specifies the audio alarm associated with ECG REPLACE
ELECTRODES.
User specifies the audio alarm associated with ECG ARTIFACT.
User specifies the color for information displayed in ECGs waveform area.
It opens the Config ECG menu.
calibration pulse
Display filter
QRS width
VTACH
Lead fail
Replace electrodes
Artifact
Config settings...
N/A
green
message
Normal
25.0 mm/s
Advanced settings...
N/A
1X
User specifies the multiplying factor used to change the appearance of the
ECG waveforms displayed.
Waveform size
Lead II
Factory Default
Lead selection
Setup ECG
Adult
Pediatric
Neonate
C-7
C-8
Advanced settings...
lo
diastolic hi
lo
mean hi
lo
Adjust limits
systolic hi
User specifies the auto mode interval for step 4 of the protocol.
User specifies the auto mode interval for step 3 of the protocol.
User specifies the auto mode interval for step 2 of the protocol.
Auto BP
repeat
4nd BP Series
repeat
3nd BP Series
repeat
2nd BP Series
repeat
User specifies the auto mode interval for step 1 of the protocol.
1st BP Series
N/A
A=40, P=40
A=140, P=100
A=30, P=30
A=120, P=90
A=80, P=70
A=200, P=150
auto-set
manual
x1
q60min
x2
q30min
x4
q15min
x4
q5min
N/A
no
When yes, the speed of the ECG waveform is fixed at 25.0 mm/s. When no,
the speed of the ECG waveform changes according to the Cardiac sweep
speed menu choice.
An extended menu where the user may configure a custom auto mode
protocol.
Factory Default
Setup NIBP
Adult
Pediatric
Neonate
Revision C
Revision C
User specifies the audio alarm associated with NIBP OVERPRESSURE.
User specifies the audio alarm associated with NIBP PUMP TIMEOUT.
User specifies the audio alarm associated with NIBP TOTAL TIMEOUT.
Overpressure
Pump timeout
Total timeout
no
yes
yes
25.0 mm/s
warning
Advanced settings...
no
N/A
hi 100 A = lo 90, P = lo
92
Adjust limits
yes
purple
It allows the user the option of viewing the SpO2 waveform area.
User specifies the color for information displayed in NIBPs vital sign area as
well as information related to NIBP that is displayed in the Limit and Full and
Mini trends windows.
procedural
procedural
procedural
procedural
procedural
N/A
warning
auto
Factory Default
View waveform?
Setup SpO2
User specifies the audio alarm associated with NIBP LEVEL TIMEOUT.
No determination
Level timeout
User specifies whether the limit alarms associated with NIBP are issued as
either warning or crisis alarms.
User specifies the pressure the Monitor initially pumps to for the next
determination.
Adult
Pediatric
Neonate
C-9
C-10
User specifies the audio alarm associated with SpO2 SENSOR FAULTY.
User specifies the audio alarm associted with SpO2 SENSOR OFF.
User specifies the audio alarm associated with SpO2 SIGNAL QUALITY
User specifies the color for the information displayed in SpO2s vital sign and
waveform areas as well as information related to SpO2 that is displayed in the
Limit and Full and Mini trends windows.
It opens the Config SpO2 menu.
User specifies the averaging time used by the Masimo SpO2 algorithm to
calculate SpO2 values
User specifies the sensitivity thresholds used by the Masimo SpO2 algorithm
for calculating SpO2 values under low perfusion conditions
User specifies the whether or not the Masimo SpO2 algorithm calculates the
SpO2 values quicker
User specifies the setting for the Nellcor SpO2 algorithm for rejecting noise
and calculating SpO2 values.
User specifies the time setting for the Nellcor SpO2 algorithm to holf off limit
alarms or OFF to disable any hold off.
Sensor faulty
Sensor off
Config settings...
lo
Adjust limits
User specifies the multiplying factor used to change the appearance of the
RESP waveform.
Waveform size
hi
1X
It allows the user the option of viewing the RESP waveform area.
View waveform?
A=6, P=10
A=30, P=60
N/A
yes
User can choose the lead from which the respiration waveform is characterized
and the impedance respiration rate is derived.
Lead II
blue
message
procedural
Lead to analyze
Setup RESP
procedural
Sensor disconnected
procedural
procedural
User specifies the audio alarm associated with SpO2 LOST PULSE.
Lost pulse
N/A
Factory Default
Adult
Pediatric
Neonate
Revision C
Revision C
Disconnected
Too Hot
Advanced settings...
User specifies the audio alarm associated with TEMP BAD PROBE
Choose mode
Bad Probe
Unit of measure
Setup TEMP
Config settings...
User specifies the color for information displayed in RESPs waveform and vital
sign areas as well as information related to RESP that appears in the Limit and
Full and Mini trends windows.
Artifact
procedural
Saturation
procedural
procedural
procedural
N/A
N/A
predictive
A=no; P=yes
N/A
blue
procedural
procedural
warning
N/A
User specifies the audio alarm associated with RESP LEAD FAIL.
warning
Lead fail
auto
The user specifies the alarm type associated with RESP RATE
APPROACHING HR.
Cardiogenic filter
A/P=12.5mm/s
N/A
Factory Default
Resp approaching
User specifies the sweep speed for all respiratory waveforms. This choice
appears in all menus associated with respiratory waveforms. A change in one
menu affects all respiratory waveforms.
Advanced settings...
Adult
Pediatric
Neonate
C-11
C-12
yellow
N/A
no
User specifies the color for information displayed in TEMPs vital sign area as
well as information related to TEMP that is displayed in the Limit and Full and
Mini trends windows
It opens the Config TEMP menu
User specifies whether the units used to display the temperature must always
fixed on C
Config settings...
setup continuous
setup timed
It allows the user to copy the contents of the active table onto the confirmed
user default table.
Config settings...
It allows the user to specify the location of printouts that result from pressing
the record .
Chart speed
setup timed
N/A
N/A
20
N/A
N/A
at bedside
25.0 mm/s
N/A
N/A
OFF
no
User specifies whether or not a Vitals Summary block of info is printed at the
beginning of real time printouts.
setup continuous
no
Print on alarm?
Setup RECORDER
Factory Default
Adult
Pediatric
Neonate
Revision C
Revision C
mm/dd/yy
military
English
yes
yes
no
no
OK
User specifies the format the date appears on the screen and on all recordings.
User specifies how time is to be formatted when it appears on the display and
all recordings. For all languages, when am/pm is confirmed, the time is
displayed as HH:MM am or HH:MM pm.
The user specifies the language to be used for text that appears throughout
the Monitors user interface. The language choice is active for all patient
settings tables. The new language becomes effective upon selection but is not
copied onto the active table until save default changes has been confirmed.
The language confirmed has no influence on service mode which is always in
English.
The user specifies whether the unit of measure for all parameters is displayed
in each parameters respective vital sign area.
The user specifies whether alarms limits for all parameters are displayed in
each parameters respective vital sign area.
The user specifies whether pressure values are displayed in kPa is in addition
to being displayed in mmHg.
When selected, the user is prompted with a yes/no dialog box with the text
CAUTION! This will reset ALL defaults! Are you sure you want to do this?
It allows the user to transmit the contents of all six tables either to another
Monitor or PC.
Language
Display units?
Display limits?
N/A
full color
Config settings...
Advanced Settings...
N/A
An extended menu where the user specifies a new date and time by manually
adjusting the month, day, year, hour, minute and second
no
User controls when battery icon is viewable when monitor in not on battery
power.
no
When selected, the user is prompted with a yes/no dialog box with the text
CAUTION! This will exit configuration mode. All unsaved changes will be lost.
Are you sure you want to do this?
N/A
Factory Default
Adult
Pediatric
Neonate
C-13
C-14
User specifies serial data transfer for this port.
Baud rate
Serial 2 setup
Startup mode
Baud rate
Startup mode
Serial 1 setup
Remote access
IP address
It allows the user to specify the host communications address for this particular
monitor.
Waveforms to send
Unit address
Config HostComm...
9600
ASCII cmd
N/A
(Viewable only in
operation
mode)38400
ASCII cmd
NA
None
2 chosen
0.0.0.0
Factory Default
Adult
Pediatric
Neonate
Revision C
D-1
Auxiliary Output
Revision C
D-2
Revision C
Auxiliary Output:
Auxiliary Output
The auxiliary output is a nonisolated signal input/output meant only for
connection to another medical device that is compliant with IEC 601-1.
Connection to a noncompliant device could result in a safety risk by exceeding
current limits.
The auxiliary output connector provides an analog ECG waveform output. The
analog ECG waveform signal is a high level (1 V/mV) representation of the
primary lead. The connector is a two-conductor 3.5 mm stereo jack that mates
with a 3.5 mm stereo plug (Shogyo SPY 1011, Shogyo SPY 1012, or equivalent).
Cable construction should be shielded, three conductor with PVC insulation or
equivalent insulating material.
Pin Out:
Tip
No Connection
Ring
Sleeve
Signal Return
Dynamic range
Scale Factor*
1 V/mV 5%
<100
Frequency Response
Delay**
22 5ms
Revision C
D-3
Auxiliary Output:
D-4
Revision C
Revision C
EMC Compliance
E-1
E-2
Revision C
WARNING
The equipment or system should not be used adjacent to, or
stacked with, other equipment. If adjacent or stacked use is
necessary, the equipment or system should be tested to verify
normal operation in the configuration in which it is being used.
Compliance
RF Emissions
EN 55011
Group 1
RF Emissions
EN 55011
Class A
Harmonic Emissions
EN 61000-3-2
Class A
Voltage Fluctuations/
Flicker Emissions
EN 61000-3-3
Complies
Revision C
E-3
Compliance Level
6 kV contact
6 kV contact
8 kV air
8 kV air
Electrical Fast
Transient/Burst
EN 61000-4-4
2 kV for power
supply lines
1 kV for input/output
lines
1 kV for input/
output lines
Surge
EN 61000-4-5
1 kV differential mode
1 kV differential
mode
2 kV common mode
2 kV common
mode
3 A/m
3 A/m
Voltage dips,
short
interruptions
and voltage
variations on
power supply
input lines
EN 61000-4-11
Power
Frequency (50/
60 Hz) Magnetic
Field
EN 61000-4-8
NOTE:
Ut is the AC mains voltage prior to application of the test level.
E-4
Revision C
Compliance
Level
Conducted RF
EN 61000-4-6
3 Vrms
150 KHz to 80 MHz
3 V rms
Radiated RF
EN 61000-4-3
3 V/m
80 MHz to 2.5 GHz
3 V/m
d = 1.2
d = 1.2
d = 2.3
Note 1: At 80 MHz and 800 MHz, the higher frequency range applies.
Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by reflection from
structures, objects, and people.
a
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land
mobile radio, AM and FM radio broadcast, and TV broadcast cannot be predicted theoretically with
accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site
survey should be considered. If the measured field strength in the location in which the equipment is used
exceeds the applicable RF compliance level above, the equipment should be observed to verify normal
operation. If abnormal performance is observed, additional measures may be necessary, such as reorienting or relocating the equipment.
b
Over the frequency range 150 KHz to 80 MHz, field strengths should be less than 3 V/m.
Revision C
E-5
0.01
0.12
0.12
0.23
0.1
0.38
0.38
0.73
1.2
1.2
2.3
10
3.8
3.8
7.3
100
12
12
23
d = 1.2
d = 2.3
At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
For transmitters rated at a maximum output power not listed above, the
recommended separation distance [d] in meters (m) can be estimated using the
equitation applicable to the frequency of the transmitter, where P is the
maximum output power rating of the transmitter in watts (W) according to the
transmitter manufacturer.
NOTE:
These guidelines may not apply in all instances. Electromagnetic
propagation is affected by absorption and reflection from structures,
objects and people.
E-6
Revision C
The table below lists cables, transducers, and other applicable accessories with
which GE Medical Systems claims EMC compliance.
NOTEAny supplied accessories that do not affect EMC compliance are not
included.
Part No
Revision C
Description
Maximum Lengths
2021141-001
2022948-001
2017003-001
3.6M
2017004-001
3.6M
900716-001
24 in.
002200
002201
002202
002203
2017002-003
2002800-001
2021406-001
70124021
2008774-001
2008775-001
615118
2.4M
3M
E-7
E-8
Revision C
Revision C
Connection
F-1
Connection:
For your notes
F-2
Revision C
Technical Specifications
Environmental Specifications
Item
Description
Operational
Temperature
5 C to 40 C (41 F to 104 F)
Operating
Atmospheric
Storage
Transportation/
Atmospheric Pressure
Storage Temperature
Humidity Range
0% to 95% non-condensing
Radio Frequency
IPX 0
Product Compatibility
CAUTIONS
Do not use the unit with any monitor except the Dash 2500 Monitor
Specified.
Contact your sales representative before connecting the network
the unit.
There are no user-serviceable parts within this device.
Revision C
When printing ECG from a Unity device, the filter setting will always be .05120 Hz, regardless of the filter setting on the Monitor.
F-3
External Connectors
The following tables provide pin-by-pin descriptions and signal names for the
external connectors.
Ethernet Port
This RJ-45 connector connects the network module to the monitoring network.
8-Pin Ethernet Connector
F-4
Pin
Name
Input/Output
Description
LAN_TRANSMIT+
LAN_TRANSMIT
LAN_RECEIVE+
NC
NC
LAN_RECEIVE
NC
NC
Revision C
Service Port
The RS-232 connector is provided as one asynchronous port for software
downloading.
8-Pin RJ-45 Connector
Pin
Revision C
Name
Input/Output
HC_232_TX
HC_232_RX
NC
COMMON
SP_232_TX
NC
SP_232_RX
NC
Description
Signal Common
O
F-5
LEDs
System Status LED Chart
LED State
F-6
Function
Description
Green ON Solid
Working
Yellow ON solid
Pending
Duplicate
address
Boot Code
Downloading
Communicatio
ns error
Other errors
Revision C
Pin Number
Revision C
Function
Ground
Fused +5 volts
No Connection
Ground
Remote Alarm
No Connection
10
No Connection
11
12
13
14
No Connection
15
No Connection
F-7
Connection: Cleaning
Cleaning
Wipe the exterior of the module with a cloth slightly dampened with mild
detergent. Wipe with a 10% solution of household bleach, Amphyl (reg.
trademark of Sterling Winthrop, Inc.) a green soap and water solution, or
Cidex (reg trademark of Johnson & Johnson).
Do not clean the module with isopropyl alcohol or other solvents, except
when preparing for mounting adhesives.
Preparation
Special Equipment
This procedure requires the installer to have the following:
F-8
1.
2.
Remove the Interface cover on the rear of the Dash 2500 Monitor.
3.
Connect the module to the 15-pin connector and insert the supplied screw
through the module. Fasten securely.
4.
Revision C
2.
Connect network cable to the port labeled as Ethernet on the rear of the
Dash 2500 Monitor. Connect the other end of the cable to the clinical
network.
3.
Set up IP address:
The Dash 2500 Monitor is an Ethernet device bounded with an IP address. It
communicates with CIC devices via the clinical network. There are two ways to
set up the IP address:
Default IP address. (If your CIC IP is 126.XXX.YYY.ZZZ, please use this method)
In this method, the Dash 2500 Monitor IP address will be set automatically.
No configuration is needed.
By factory default, the IP address in Dash 2500 Monitor configuration menu
is set as 0.0.0.0. The monitor will automatically generate a default IP
address calculated from its MAC address. This IP address is unique in all GE
Unity devices and follows below formula:
126. XXX. YYY. ZZZ
where the XXX, YYY and ZZZ are the last three numbers of the monitor MAC
address.
The IP address of the Dash 2500 Monitor must be in the same network
address of the intended CIC devices. For example, if the CIC IP address is
172.16.***.*** in a class-B sub-network, the valid IP address range for
monitor is 172.16.0.1~172.16.255.254.
Revision C
2.
Turn and press the Trim Knob to select other system settings.
3.
F-9
7.
8.
Re-set up
To re-set up type for the network function, enter into boot service menus to clear
config memory. For detail steps, refer to Using the Boot Service Menu.
F-10
Revision C
Preparation
Special Equipment
This procedure requires the installer to have the following:
Windows XP compatible
CD drive
Interconnection
To connect a personal computer (PC) to the network module (unit), complete the
following steps in the order given.
1.
Connect the RJ-45 cable to the RJ-45 connector labeled Serial 2 on the unit.
2.
Connect the other end of the RJ-45 cable to PC DIDCA, which is connected
to the 9-pin D-type connector labeled COMMS or COMM 1 at the rear of the
PC.
3.
Connect power cable between the AC connector and the AC power outlet.
PC Setup
Follow the steps in the order listed.
1.
2.
3.
Revision C
F-11
While holding down the lowercase b key continuously at the PC, power ON
the Unity Adapter (Dash 2500 Monitor).
Each selection in the BOOT SERVICE MENU is described below. If you enter the
wrong configuration data for any selection, exit the menu, reenter the menu,
and type over the old data.
Select 1: Change Ethernet Address. This selection allows you to change the
Ethernet address of the unit.
Select 2: Change Internet Address. This selection allows you to change the
Internet address of the unit. To generate an Internet address from an
Ethernet address, GE Medical Systems Information Technologies simply
converts the last three bytes of the Ethernet address from hex to decimal.
This combines the unique identifier along with the network ID to produce
the entire Internet address.
NOTE:
This IP address will be overwritten by a valid address from the Dash
2500 Monitor when it is connected.
WARNING
Duplication of an Internet address on the network will cause
lost data. If you change the factory assigned Internet address,
you must first record all other Internet addresses used on your
network to avoid duplication.
F-12
Revision C
Select 3: Change the Network Module ID Number. This selection allows you
to input the unique ID number from the ID number label on the unit. This
number is coded into the software at the factory when the ID number label
is applied. Ensure that the ID number from the label is the same as the
number you input.
New IPC-1928 ID: ___ (range 0 to 9999)(Press ENTER key.)
The IPC-1928 ID has been set to ___.
Revision C
Select 4: Clear Config Memory. This selection allows you to clear the
configurable memory and ID number. The Ethernet and internet addresses
are not affected when you clear the configuration.
Select 5: Country Selection (xxx, xxx is the current country). This selection
selects what country and software the unit is using. The default is English.
F-13
Service Menus
The service menus described in this section are presented for advanced users
and service personnel only. In most cases, access to these menus will not be
necessary and only experienced technicians should attempt to use them.
Preparation
Special Equipment
This procedure requires the installer to have the following:
Windows XP compatible
CD drive
Interconnection
To access the service menu, complete the following steps in the order given.
1.
Connect the RJ-45 cable to the RJ-45 connector labeled Serial 2 on the unit.
2.
Connect the other end of the RJ-45 cable to PC DIDCA, which is connected
to the 9-pin D-type connector labeled COMMS or COMM 1 at the rear of the
PC.
PC Setup
Follow the steps in the order listed.
1.
2.
3.
4.
F-14
Revision C
NOTE:
No password prompt will appear on screen
Type password: service (all lower case)
Press Enter key
Communication Tips
Use the following controls if you have problems communicating between the
unit and the PC.
If you enter the wrong configuration data, just exit the menu, reenter the
menu, and type over the old data.
Select 1: Revisions. This selection lists the revision of the network module
main code.
Select 2: Error Log. This selection lists all errors recorded by the network
module main code, similar to the example below.
1 View Next
Revision C
F-15
Select 3: Event Log. This selection lists all events recorded by the network
module main code, similar to the example below.
Enn
WHEN
EVENT
ID
DATA
DESCRIPTION
0000
01/27/06
13:00:27
Cable Off
0001
01/27/06
13:00:27
Cable Off
0002
01/27/06
13:02:39
30203
Cable On
0003
01/27/06
13:02:46
50106
Cable On
0004
01/27/06
13:02:55
10101
Cable On
1 View Next
2 View Previous
3 Clear Error Log
4 Exit
Enn = the event number
WHEN = date and time1 View Next
WHEN = date and time
F-16
Revision C
Select 7: Rwhat Database. This selection allows the user to see all the Rwhat
packets on the network in different formats.
RWHAT SERVICE MENU
1 Display All
2 Display All (Abbreviated)
3 Display for Resource
4 Display for Resource (Abbreviated)
5 Search for String
6 Search for String (Abbreviated)
7 Exit
Select 8: Set the Time and Date. In this version the Time/Data setting
funtion is disabled.
TIME AND DATE SERVICE MENU
13:08:09 27-JAN-2006
Setting of Time/Date has been disabled in this software
version
Revision C
Select 10. View Parameters. This selection allows the user to view the
parameter values in a demonstration mode.
F-17
F-18
Revision C
World Headquarters
European Representative
Asia Headquarters
GE Medical Systems
Information Technologies, Inc.
8200 West Tower Avenue
Milwaukee, WI 53223 USA
Tel:+ 1 414 355 5000
1 800 558 5120 (US only)
Fax:+ 1 414 355 3790
GE Medical Systems
Information Technologies GmbH
Munzinger Strae 3-5
D-79111 Freiburg
Germany
Tel: + 49 761 45 43 - 0
Fax: + 49 761 45 43 - 233
GE Medical Systems
Information Technologies Asia; GE (China) Co., Ltd.
GE China Technology Park
1 Huatuo Road
Shanghai 201203, P.R. China
Tel: + 86 21 3877 7888
Fax: + 86 21 3877 7451
0459