Mks 626a
Mks 626a
Mks 626a
Rev G, 10/01
ZERO
ZERO
Type 622
Pressure
Transducer
+15 VDC-35 mA
PIN SIGNAL
1
Pwr Com
2
Sig Com
3
Sig Output
4
-15 VDC
5
+15 VDC
Chassis
Type 62x
Pressure
Transducer
+ VDC 250 mA
PIN
2
5
6
7
12
15
SIGNAL
Sig Output
Pwr Com
-15 VDC
+15 VDC
Sig Com
Chassis
Table of Contents
Table of Contents
iii
Table of Contents
iv
Figures
Figure 1: Preferred Method To Connect an Overall Metal Braided Shielded Cable................. 22
Figure 2: Alternate Method To Connect an Overall Metal Braided Shielded Cable................. 22
Figure 3: Metal Single and Double Ferrule Compression-type Vacuum Fittings ..................... 24
Figure 4: Location of the Zero Pot on Transducer Endcaps ...................................................... 26
Figure 5: Location of the COARSE ZERO Adjustment Control............................................... 27
Figure 6: Dimensions of the Types 622, 626, 627, 628, and 629 Transducers.......................... 29
Figure 7: Power, Signal, and Chassis Grounding Scheme......................................................... 30
Figure 8: Using Pin 7 as an External Zero Adjustment ............................................................. 37
Figure 9: Type 629 Transducers Connected to Reject Common Mode Voltages ..................... 38
Tables
Table 1: Definition of Symbols Found on the Unit ........................................................................2
Tabelle 2: Bedeutung der am Gert angebrachten Symbole...........................................................6
Tableau 3: Dfinition des symboles apparaissant sur l'unit ........................................................10
Tabla 4: Definicin de los smbolos hallados en la unidad ...........................................................14
Table 5: Interface Cables ..............................................................................................................20
Table 6: Highest Pressure Suggested for Proper Zero Adjustment ..............................................26
Table 7: Required Inputs for Types 622, 626, 627, 628, and 629 Transducers............................30
Table 8: Pinout of Type 622 Transducer ......................................................................................31
Table 9: Pinout of Types 626, 627, and 628 Transducers ............................................................31
Table 10: Pinout of Type 629 Transducer ....................................................................................32
Table 11: Suggested Pressures for Reading and Control..............................................................36
Table 12: Troubleshooting Chart ..................................................................................................40
vi
Definitions of WARNING, CAUTION, and NOTE messages used throughout the manual.
Warning
Caution
Note
|
Earth (ground)
IEC 417, No.5017
Protective earth
(ground)
IEC 417, No.5019
Equipotentiality
IEC 417, No.5021
Direct current
IEC 417, No.5031
Alternating current
IEC 417, No.5032
Class ll equipment
IEC 417, No.5172-a
Three phase
alternating current
IEC 617-2 No.020206
Caution, refer to
accompanying
documents
ISO 3864, No.B.3.1
Caution, risk of
electric shock
ISO 3864, No.B.3.6
On (Supply)
IEC 417, No.5007
Frame or chassis
IEC 417, No.5020
Off (Supply)
IEC 417, No.5008
Warnung!
Vorsicht!
Das Symbol VORSICHT! weist auf eine Gefahr fr das Gert hin. Es
macht auf einen Bedienungsablauf, eine Arbeitsweise oder eine
sonstige Gegebenheit aufmerksam, deren unsachgeme Ausfhrung
bzw. ungengende Bercksichtigung zu einer Beschdigung oder
Zerstrung des Gertes oder von Teilen des Gertes fhren kann.
Hinweis
|
Ein (Energie)
IEC 417, No.5007
Aus (Energie)
IEC 417, No.5008
Erdanschlu
IEC 417, No.5017
Schutzleiteranschlu
IEC 417, No.5019
Masseanschlu
IEC 417, No.5020
Aquipotentialanschlu
IEC 417, No.5021
Gleichstrom
IEC 417, No.5031
Wechselstrom
IEC 417, No.5032
Gleich- oder
Wechselstrom
IEC 417, No.5033-a
Durchgngige
doppelte oder
verstrkte Isolierung
IEC 417, No.5172-a
Warnung vor
gefhrlicher
elektrischer Spannung
ISO 3864, No.B.3.6
Hhere Temperatur
an leicht
zugnglichen Teilen
IEC 417, No.5041
DreileiterWechselstrom
(Drehstrom)
Avertissement
Attention
Remarque
|
Marche
(sous tension)
IEC 417, No.5007
Terre (masse)
IEC 417, No.5017
Terre de protection
(masse)
IEC 417, No.5019
Masse
IEC 417, No.5020
Equipotentialit
IEC 417, No.5021
Courant continu
IEC 417, No.5031
Courant alternatif
IEC 417, No.5032
Courant continu et
alternatif
IEC 417, No.5033-a
Matriel de classe II
IEC 417, No.5172-a
Courant alternatif
triphas
IEC 617-2, No.020206
Attention : se reporter
la documentation
ISO 3864, No.B.3.1
Attention : risque de
choc lectrique
ISO 3864, No.B.3.6
Attention : surface
brlante
IEC 417, No.5041
10
11
12
Advertencia
Precaucin
Nota
13
|
Encendido
(alimentacin elctrica)
IEC 417, N 5007
Apagado
(alimentacin elctrica)
IEC 417, N 5008
Puesta a tierra
IEC 417, N 5017
Proteccin a tierra
IEC 417, N 5019
Caja o chasis
IEC 417, N 5020
Equipotencialidad
IEC 417, N 5021
Corriente continua
IEC 417, N 5031
Corriente alterna
IEC 417, N 5032
Corriente continua y
alterna
IEC 417, N 5033-a
Equipo de clase II
IEC 417, N 5172-a
Corriente alterna
trifsica
IEC 617-2, N 020206
Precaucin. Consulte
los documentos
adjuntos
ISO 3864, N B.3.1
Precaucin. Riesgo
de descarga elctrica
ISO 3864, N B.3.6
Precaucin. Superficie
caliente
IEC 417, N 5041
14
15
16
Introduction
Introduction
The MKS Baratron Type 622A, 626A, 627A, 628A, and 629A Absolute Pressure Transducers
are part of the MKS family of general purpose pressure transducers designed to provide accurate,
reliable, and repeatable pressure measurements in the range from 1K Torr to as low as 0.05 Torr
F.S. (627 and 629). These instruments operate with 15 VDC (5%) input @ 35 mA, 250 mA,
or 500 mA, depending upon the unit, and provide 0 to 10 VDC output linear with pressure. All
instruments expose only Inconel to the process permitting use with corrosive or dirty gases and
eliminating contamination of the process with transducer materials. Measurements are
independent of gas composition and all of the units have a minimum measuring range of four
decades.
Using the latest single-sided, dual-electrode Inconel transducer design, coupled with a low
impedance, fixed-frequency bridge signal conditioner, these instruments are capable of
withstanding high overpressure conditions (45 psia) with minimal or no shifts in output over their
range. The advanced bridge signal conditioning technology provides high accuracy and
operation which is extremely temperature-stable at operating pressure.
Protection from RF interference and noisy electrical environments is increased by the use of a
metal case, by internal design elements, and by the use of surge and ESD suppression networks
and RFI filtering on all inputs and outputs.
The Types 622 and 626 transducers have an accuracy of 0.25% of reading, with the exception of
the 626A 0.10T which has an accuracy of 0.50% of reading (optional 0.15% of reading, 10Torr
to 1000 Torr only). The units are identical except for their connectors; the 622 unit has a 5-pin
terminal block connector and the 626 unit has a 15-pin Type D connector. These instruments
have been developed for applications requiring a small, economical transducer, which delivers
accurate and repeatable measurements with a resolution of 1 x 10-4 F.S.
The Types 627, 628, and 629 instruments are designed specifically to meet the needs of vacuum
process systems where environmental and process conditions are particularly demanding.
Temperature control of the transducer minimizes the effects of ambient or process temperature
variations typically encountered in process line environments. The 627 and 629 transducers
control the transducer temperature at 45 C and the 628 unit controls the transducer temperature
at 100 C, thereby minimizing contamination from the process. The Type 627 and 629
transducers have an accuracy of 0.12% of reading (0.1 and 0.05 Torr units have an accuracy of
0.15% of reading), and the Type 628 transducer has an accuracy of 0.25% of reading.
17
Customer Support
Standard maintenance and repair services are available at all of our regional MKS Calibration
and Service Centers, listed on the back cover. In addition, MKS accepts the instruments of other
manufacturers for recalibration using the Primary and Transfer Standard calibration equipment
located at all of our regional service centers. Should any difficulties arise in the use of your
Type 622, 626, 627, 628, or 629 instrument, or to obtain information about companion products
MKS offers, contact any authorized MKS Calibration and Service Center. If it is necessary to
return the instrument to MKS, please obtain an ERA Number (Equipment Return Authorization
Number) from the MKS Calibration and Service Center before shipping. The ERA Number
expedites handling and ensures proper servicing of your instrument.
Please refer to the inside of the back cover of this manual for a list of MKS Calibration and
Service Centers.
Warning
18
How To Unpack
How To Unpack
MKS has carefully packed the Type 622/626/627/628/629 unit so that it will reach you in perfect
operating order. Upon receiving the unit, however, you should check for defects, cracks, broken
connectors, etc., to be certain that damage has not occurred during shipment.
Note
Do not discard any packing materials until you have completed your
inspection and are sure the unit arrived safely.
If you find any damage, notify your carrier and MKS immediately. If it is necessary to return the
unit to MKS, obtain an ERA Number (Equipment Return Authorization Number) from the MKS
Service Center before shipping. Please refer to the inside of the back cover of this manual for a
list of MKS Calibration and Service Centers.
Unpacking Checklist
Standard Equipment:
The Type 622, 626, 627, 628, or 629 Absolute Pressure Transducer (specified at time of
order).
The Type 622 transducer comes with a mating connector for the transducers 5-pin terminal
block.
Optional Equipment:
The Types 622, 626, 627, 628, and 629 transducers are compatible with:
Most MKS Pressure, Flow, Flow Ratio, and Throttling Valve Controllers
Optional RM.-6 rack mount kit permits mounting one or two Readouts and/or Controllers in
a 19 rack
Electrical Connector Accessories Kit, 62XA-K1 (where 62X is either 626, 627, 628, or
629)includes a mate for the electrical connector
19
Interface Cables
Interface Cables
As of January 1, 1996, most products shipped to the European Community must comply with the
EMC Directive 89/336/EEC, which covers radio frequency emissions and immunity tests. In
addition, as of January 1, 1997, some products shipped to the European Community must also
comply with the Product Safety Directive 92/59/EEC and Low Voltage Directive 73/23/EEC,
which cover general safety practices for design and workmanship. MKS products that meet
these requirements are identified by application of the CE Mark.
To ensure compliance with EMC Directive 89/336/EEC, an overall metal braided shielded cable,
properly grounded at both ends, is required during use. No additional installation requirements
are necessary to ensure compliance with Directives 92/59/EEC and 73/23/EEC.
Note
Interface Cables
Product
622
CB473-1-10
CB112-2-10
626
CB127-1-10
CB259-5-10
627
CB127-1-10
CB259-5-10
628
CB127-1-10
CB259-5-10
629
20
Interface Cables
21
Interface Cables
Transducer
Transducer Housing
Transducer
Overall Insulation
(if present)
Transducer Housing
22
Ambient operating temperature should remain between 0 C and 50 C (32 F and 122 F)
626 0.01 Torr ambient operating temperature should remain between 15C and 40C (59F
and 104F)
Input required is 15 VDC (5%) @ 35 mA (maximum) with <20 mV p-p noise and ripple
Ambient operating temperature should remain between 15 C and 40 C (59 F and 104 F)
Input required is 15 VDC (5%) @ 250 mA (maximum) with <20 mV p-p noise and ripple
Type 628:
Ambient operating temperature should remain between 20 C and 70 C (68 F and 158 F)
Input required is 15 VDC (5%) @ 500 mA (maximum) with <20 mV p-p noise and ripple
For additional product requirements refer to Appendix A: Product Specifications, page 41.
23
Fittings
Fittings
The transducer port will easily carry the weight of the transducer. The following is a list of
available fittings:
NW-16-KF
Mini-CF
Caution
MKS does not warranty the 622, 626, 627, 628, or 629 transducer
when single or double metal ferrule compression-type vacuum
fittings (shown in Figure 3, page 24) are used because damage will
occur to the transducer when improper tightening procedures are
followed.
Double Ferrule
Single Ferrule
24
Setup
Setup
Mount the transducer with the inlet port pointing (vertically) downward. Although the unit can
be mounted in any orientation, mounting it as suggested allows any foreign matter entering the
pressure port to fall away from the diaphragm.
Isolate the unit from vibration as much as possible. When not subject to gas damping at low
pressure, the diaphragm may become susceptible to resonance. The 0.05, 0.1 and 1 Torr F.S.
units are most sensitive and you should isolate these units from any vibration that exists.
Remember to isolate the vibration through the cable as well as through the port.
Checking the Transducer Zero
Check the transducer zero prior to the initial operation and then periodically as required. The
zero can be set (or reset) by adjusting the zero potentiometer located on the top cover of the
transducer or at the front panel of any MKS Power Supply/Readout which is being used. The
unit must be fully stabilized before you change the zero adjustment. Stabilization is
accomplished by allowing the unit to warm up for approximately:
2 to 4 hours for the Types 627, 628, and 629 instruments ( 1 Torr)
8 hours for the 0.05 Torr (627 and 629) and 0.1 Torr (627, 628, 629)
Pump the unit, with the power on, down to a pressure less than the transducers resolution
(0.01% of Full Scale).
Note
Once the unit is properly warmed up and stabilized, the zero adjustment
must be made at a pressure less than the transducers resolution (0.01%
of F.S.). Low range transducers (0.1 and 0.05 Torr) should be pumped
for at least one hour after exposure to air to remove any moisture and to
allow the pressure to stabilize.
Zeroing a transducer above its stated minimum resolution creates a zero offset relative (or
unique) to the system in which the transducer is located. All subsequent readings are then linear
and accurate relative to the offset value.
Refer to Figure 4, page 26, for the location of the zero potentiometer on the Types 622, 626, 627,
628, and 629 transducers.
Note
If available pressures are not sufficiently low to set the transducer zero,
you may use a vacuum leak detector with sufficient vacuum pumping (to
achieve proper zeroing pressures). In this case, mount the transducer on
the leak detector in the same plane of orientation as it will be during
actual use.
25
Setup
ZERO Potentiometer
ZERO
ZERO
Type 622
Pressure
Transducer
+15 VDC-35 mA
PIN SIGNAL
1
Pwr Com
2
Sig Com
3
Sig Output
4
-15 VDC
5
+15 VDC
Chassis
Type 626
Pressure
Transducer
+15 VDC-35 mA
ZERO
PIN
2
5
6
7
12
15
SIGNAL
Sig Output
Pwr Com
-15 VDC
+15 VDC
Sig Com
Chassis
Type 62X
Pressure
Transducer
+15 VDC-250 mA
PIN
2
5
6
7
12
15
SIGNAL
Sig Output
Pwr Com
-15 VDC
+15 VDC
Sig Com
Chassis
BARATRON
BARATRON
Type 622
Type 626
3. Pump the unit down to a pressure below its resolution (refer to Table 6 for recommended
pressure levels).
4. Using a small screwdriver, adjust the ZERO pot until the readout displays zero (0000).
Highest Pressure Suggested for Proper Zero Adjustment
Full Scale Range (Torr)
0.05*
0.1**
< 5 x 10-6
< 5 x 10-5
< 1 x 10-4
10
< 5 x 10-4
100
< 5 x 10-3
1K
< 5 x 10-2
26
Setup
Note
Use the coarse zero adjustment only if the zero potentiometer fails to
provide sufficient adjustment.
The COARSE ZERO control is accessible through the side of the instrument, as shown in
Figure 5.
ZERO
Potentiometer
COARSE ZERO
Adjustment
ZERO
Type 62X
Pressure
Transducer
+15 VDC-250 mA
PIN
2
5
6
7
12
15
SIGNAL
Sig Output
Pwr Com
-15 VDC
+15 VDC
Sig Com
Chassis
27
Setup
28
Setup
Dimensions
Note
2.56 DIA
(65.0)
2.56 DIA
(65.0)
0.31
(7.9)
0.13
(3.30)
3.17
(80.52)
3.17
(80.52)
4.75
(120.65)
4.93
(125.22)
1.45
(36.83)
1.45
(36.83)
0.500 DIA
(12.70)
0.500 DIA
(12.70)
Type 622A
Type 626A
3.00 DIA
(76.2)
0.13
(3.30)
5.19
(131.8)
3.76
(95.50)
Type 627A and 628A Units
1.30
(33.02)
0.500 DIA
(12.7)
Figure 6: Dimensions of the Types 622, 626, 627, 628, and 629 Transducers
29
Electrical Information
Electrical Information
The Types 622, 626, 627, 628, and 629 transducers require an external power source capable of
supplying the voltages stated in Table 7. Noise and ripple should be less than 20 mV p-p. You
may use any readout device which has input capabilities of less than 0 to greater than 10 VDC,
and impedance greater than 10K ohms.
Note
The ground of any external power supply and readout should be the same
as the transducer ground (chassis ground) to minimize any possible
ground loops which can affect the performance and stability of the
system.
Required Inputs for Types 622, 626, 627, 628, and 629 Transducers
Input Required
Type 628
At Startup
15 VDC (5%) @
35 mA
15 VDC (5%) @
250 mA
15 VDC (5%) @
500 mA
After 1 hour of
operation at 25 C
15 VDC (5%) @
35 mA
15 VDC (5%) @
150 mA
15 VDC (5%) @
350 mA
Table 7: Required Inputs for Types 622, 626, 627, 628, and 629 Transducers
Multiple Ground
Tie-Point
0 - 10 V Pressure Output
Signal Common
+
-
Readout
+ 15 V
- 15 V
+ 15V
10K
1 F
Power
Supply
Power Common
30
Electrical Information
Connectors
The 622 unit has a 5-pin terminal block connector, whose pinout is listed in Table 8. The 626,
627, and 628 units use a common 15-pin Type D connector, refer to Table 9 for the pinout.
The 629 transducer uses a 9-pin Type D whose pinout is listed in Table 10, page 32.
Assignment
Power Common
Signal Common
Signal Output
- 15 VDC
+ 15 VDC
Chassis Ground is made by using the screw labeled CHASSIS on the endcap
Table 8: Pinout of Type 622 Transducer
Pinout of Types 626, 627, and 628 Transducers
Pin Number
Assignment
Signal Output
Power Common
- 15 VDC
+ 15 VDC
12
Signal Common
15
Chassis Ground
31
Electrical Information
Assignment
Signal Output
+ 15 VDC
- 15 VDC
External Zero
Signal Common
Power Common
Table 10: Pinout of Type 629 Transducer
32
General
General
A complete pressure transducer system requires three components to convert pressure to a linear
DC voltage output: a sensor, signal conditioner, and power supply. An analog or digital meter is
required to display the DC output in pressure units.
MKS Types 622, 626, 627, 628, and 629 transducers contain two of the above components: the
sensor and signal conditioner. An MKS or MKS-compatible power supply is required to
complete the pressure to DC voltage output conversion, and an MKS or MKS-compatible display
unit is required for direct pressure readout.
Sensor
The variable capacitance sensor consists of a pressure inlet tube (port) connected to a small
chamber in the transducer body. One wall of this chamber is an elastic metal diaphragm. The
front side of the diaphragm is exposed to the gas whose pressure is to be measured. The back, or
reference side of the diaphragm faces a rigidly mounted ceramic disc containing two electrodes.
The reference side is permanently evacuated (10-7 Torr) and its vacuum is maintained with a
chemical getter system.
The diaphragm deflects with changing absolute pressure (force per unit area) independently of
the gas type or composition of the measured gas. This deflection causes an imbalance of the
sensor electrode capacitances since the distance to the diaphragm is now different for each
electrode. The imbalance of capacitances is converted to a DC voltage in the bridge. This bridge
is excited by a precision constant frequency oscillator. The resultant signal is then linearized,
zeroed, and amplified via the signal conditioner electronics, to produce a precise 0 to 10 VDC
signal scaled to the range of the transducer.
In the heated models (Types 627, 628, and 629 instruments), zero and span stability is further
increased because the sensor and bridge electronics are temperature controlled. This thermal
enclosure reduces the effects of ambient temperature changes by a factor of at least 35 (that is
35 C change in ambient will produce less than 1 C change inside the thermal housing).
Signal Conditioner/Electronics
The signal conditioner contains state-of-the-art, low impedance balanced bridge circuitry, selfcompensated for thermal stability with ambient temperature changes. Output is a DC voltage
which is linear with pressure. The transducer is then calibrated against a pressure standard to
provide a 0 to 10 Volt DC output over the range of the transducer.
33
The Type 629 transducer output is fully differential with respect to the voltage input to
the external zero (pin 7)
For details, refer to Chapter Four, Operation, page 17.
34
General
35
General
0.05*
2.5 x 10-5
2.5 x 10-4
0.1 **
5 x 10-5
5 x 10-4
5 x 10-4
5 x 10-3
10
5 x 10-3
5 x 10-2
100
5 x 10-2
5 x 10-1
1K
5 x 10-1
36
+
Type 629
V1
Voltage at
zero pressure
0V
+
V
- V1
out
Note
If the external zero pin is not being used, connect it to Signal Common
(Pin 8).
37
1 Signal Output
629 Transducer
(top view)
+ 15V
7 External Zero
1
2
8 Signal Common
3 MUX
9 Power Common
10K
1 F
Data
Acquisition
System
4 + 15V
- 15V
5
D/A Ground
Chassis
Ground
Vcm
=+1V
max
629 Transducer
(top view)
1 Signal Output
7 External Zero
8 Signal Common
+ 15V
9 Power Common
10K
1 F
4
5
+ 15V
- 15V
Chassis
Ground
629 Transducer
(top view)
1 Signal Output
7 External Zero
8 Signal Common
+ 15V
9 Power Common
10K
1 F
+ 15V
+ 15V
Power
Supply
P/S Ground
- 15V
5 - 15V
Chassis
Ground
38
General
General
In general, no maintenance is required other than proper installation and operation, and an
occasional zero adjustment. If a transducer fails to operate properly upon receipt, check for
shipping damage, and check the power/signal cable for correct continuity. Any damage should
be reported to the carrier and MKS Instruments immediately. If there is no obvious damage and
the continuity is correct, obtain an ERA Number (Equipment Return Authorization Number)
before returning the unit to MKS Instruments for service.
In production operations such as semiconductor manufacturing, verify the transducer zero (and
adjust if necessary) each time the equipment is shut down for routine maintenance.
Note
The ZERO adjustment is the only adjustment that can be made in the field.
Return the transducer to MKS Instruments for other adjustments,
calibration, or servicing.
Periodically check for wear on the cables and inspect the enclosure for visible signs of damage.
Zero Adjustment
All pressure transducers require initial and periodic zero adjustments. Make these adjustments at
a pressure lower than the transducers minimum resolution to achieve the full dynamic range
specified for the transducer. Refer to Chapter Two: Installation, page 19, for instructions on
adjusting the zero setting.
39
Troubleshooting Chart
Troubleshooting Chart
Symptom
Possible Cause
Solution
Overrange positive or
negative signal
A shorted transducer or a
damaged interconnect cable
(transducer to electronics
module).
Refer to Appendix A:
Product Specifications, page
41, and be sure the ambient
temperature is within product
requirements.
40
0 to 50 C (32 to 122 F)
CE Compliance
Electromagnetic Compatibility3
Connector
622
626
Fittings
Standard
Optional
Input required
15 VDC (5%) @ 35 mA
Inconel
Output
Temperature coefficients
Zero
Span
0.005% F.S./C4
0.04% Reading/C
Time constant
< 20 msec
1 x 10-4 F.S.
6.3 cc
Due to continuing research and development activities, these specifications are subject to change without notice.
41
0.12% of Reading
0.15% for 0.1 Torr units
15 to 40 C (59 to 104 F)
CE Compliance
Electromagnetic Compatibility6
Connector
627
629
15-pin Type D
9-pin Type D
Fittings
Standard
Optional
Input required
Inconel
Output
Temperature coefficients
Zero
Span
0.002% F.S./C7
0.02% Reading/C
Time constant
< 20 msec
1 x 10-4 F.S.
6.3 cc
Due to continuing research and development activities, these specifications are subject to change without notice.
5Do
6
not operate the Type 627 and 629 transducers outside this temperature range.
An overall metal braided shielded cable, properly grounded at both ends, is required during use.
Type 627 and 629 0.1 Torr transducers, the zero temp. coef. is 0.005% F.S./C.
7For
42
0.15% of Reading
15 to 40 C (59 to 104 F)
CE Compliance
Electromagnetic Compatibility8
Connector
627
15-pin Type D
629
9-pin Type D
Fittings
Standard
Optional
Input required
Inconel
Output
0.05
Temperature coefficients
Zero
Span
0.015% F.S./C
0.02% Reading/C
Time constant
< 20 msec
2 x 10-4 F.S.
6.3 cc
Due to continuing research and development activities, these specifications are subject to change without notice.
43
0.25% of Reading
20 to 70 C (68 to 158 F)
CE Compliance
Electromagnetic Compatibility10
Connector
Fittings
Standard
Optional
Input required
Inconel
Output
Temperature coefficients
Zero
Span
0.002% F.S./C11
0.02% Reading/C
Time constant
< 20 msec
1 x 10-4 F.S.
6.3 cc
Due to continuing research and development activities, these specifications are subject to change without notice.
8 An overall metal braided shielded cable, properly grounded at both ends, is required during use.
9Do not operate the Type 628 transducer outside this temperature range.
10 An overall metal braided shielded cable, properly grounded at both ends, is required during use.
11For Type 628 0.1 Torr transducers, the zero temp. coef. is 0.005% F.S./C.
44
Model Code
Model Code
The options of your unit are identified in the model code. The model code is identified
as follows:
62XAXXXYYZ
where:
Type 62XA
XXX
Type Number
Full Scale Range
Fittings
Accuracy
45
Model Code
Ordering Code
0.05*
0.1**
1
2
10
100
1000
U5T
.1T
01T
02T
11T
12T
13T
*Available on 627 and 629 units only
** Available on 626, 627, 628, and 629 units only
Fittings (Y)
The choice of fittings is designated by a single letter code.
Fittings
Ordering Code
A
B
C
D
E
Accuracy (Z)
The accuracy is designated by a single number code.
Accuracy
Ordering Code
0.15% of Reading,
optional for Type 622 and 626 (10, 100 and 1000 Torr only)
standard for Type 627 and 629 (0.05 and 0.1 Torr only)
0.5% of Reading, standard for Types 626 and 628 (0.1 Torr
only)
46
Index
Index
50m Torr transducers, 25
Signal Conditioner/Electronics, 15
Specifications, 6, 23-26
C
Coarse Zero adjustment, 9-10
Companion products, 5
Troubleshooting, 22
Customer support, 4
Dimensions, 11
50m Torr, 25
Electrical installation, 12
Fittings, 6
U
I
Manual organization, 4
Messages, definitions of, 2
P
Pressure
lowest for control, 17
lowest for reading, 17
R
Returning the product, 4, 5, 21
S
Safety procedures and precautions, 1-2
Sensor, 15
Setup, 7
47