Technical Information Ousaf22
Technical Information Ousaf22
Technical Information Ousaf22
16
71316931
Products
Solutions
Services
Technical Information
OUSAF22
Optical sensor combined with the OUA260 flow
assembly for the measurement of color
concentrations
Application
The sensor is used to measure color concentrations in the range of the visible
electromagnetic spectrum.
Color scale measurement
APHA/Hazen, EBC, ASBC, ASTM, ICUMSA
Measurement of color concentration
Outgoing goods inspection/purity monitoring
Color assurance
Discoloration control
Distillation monitoring
Your benefits
As slightest color changes are detected, production quality is guaranteed
Measuring range up to 2.5 AU or 50 OD (depending on optical path length)
Can be configured to measure color concentration at discrete wavelengths in the
visible regionregion
Outstanding filter properties for highest linearity
Direct concurrence with laboratory values
Integrated reference detector for the compensation of particles, bubbles and
lamp aging
Incandescent lamp for long service life and stable measured values
Cost-effective, time-efficient verification (liquid-free)
FM- and ATEX-approved lamps for applications in the hazardous area
Deployable and durable in numerous applications:
Variety of materials and process connections
Hygienic design for high degree of product safety: certified materials and CIP/SIPresistance
Can be adapted to process requirements:
Optional air purge ports to prevent the formation of condensate on the optical
windows
OUSAF22
Light absorption
The measuring principle is based on the Lambert-Beer law.
There is a linear dependency between the absorption of light and the concentration of the absorbing
substance:
A = -logAm = . c . OPL
A ... absorption, Am ... absorption measured by detector
... Extinction coefficient
c ... Concentration
OPL ... Optical path length
A light source emits radiation through the medium and the incident radiation is measured on the
detector side.
A beam splitter splits the light into two beams. One beam is used for measuring while the other acts
as a reference to compensate for particles, bubbles and lamp aging.
Once the light has passed through a filter for wavelength selection, the intensity of the light is
determined by a photodiode and converted to a photocurrent.
The subsequent conversion to absorbance units (AU, OD) is performed in the associated transmitter.
AR
OPL
Am
4
A0029408
1
1
2
3
4
5
6
7
8
Measuring system
Endress+Hauser
OUSAF22
pipe
Transmitter CM44P
Sensor: detector
Input
Measured variable
process absorption
Measuring range
0 to 2.5 AU
Max. 50 OD (depending on the optical path length)
Wavelength
Measurement 400 nm, 420 nm, 430 nm, 490 nm, 520 nm
Reference 720 nm
Power supply
Electrical connection
Endress+Hauser
The sensor is connected to the transmitter using the pre-terminated or labeled cable set CUK80 (for
connection to CM44P) or OUK20 (for connection to CVM40). The terminals and labeling may vary
depending on the transmitter in use. The cable set must be ordered separately.
OUSAF22
GY
RD
BK
WH
GN
GY
YE
YE
BK
BK
A0028384
3
A
B
Cable length
Lamp voltage
CM44P terminal
CVM40 terminal
Cable color
Assignment
P+
V1.1
YE (thick)
Lamp voltage +
S+
V1.3
YE (thin)
P-
V1.2
BK (thick)
Lamp voltage -
S-
V1.4
BK (thin)
A (1)
S1.1
RD
C(1)
S1.2
BK
SH (1)
S1.S
GY
Shield
A (2)
S2.1
WH
Sensor reference +
C(2)
S2.2
GN
Sensor reference -
SH (2)
S2.S
GY
Shield
Lamp type
OUSAF22-xxxxx
4.9 0.1
Endress+Hauser
12.6 (0.5)
90 (3.54)
OUSAF22
105 (4.13)
The safety barrier may only have a very low leak current since the optical signals from the
sensor can be in the nanoampere range. Therefore, the sensor cable shield is connected to the
ground terminal of the barrier.
On delivery, the CUK80 detector cable is permanently wired to the safety barriers. All you have to do
is simply connect the individual cable ends to the detector and transmitter.
Photometer
(Det.)
2x MTL7760ac
BK
CM44P
BK
RD
RD
GN
WH
SH (1)
GN
WH
4 GY
C (1)
A (1)
GY
C (2)
A (2)
SH (1)
Endress+Hauser
OUSAF22
EXP-1
CM44P
ATEX II 2G Ex db IIC T5 Gb
FM Cl.1, Div. 1, Groups B, C, D
YE
YE
BK
BK
P+
S+
SP-
PE
27
30
A0029440
Installation
Installation instructions
A
A0028250
6
A
B
C
D
E
Mounting angles. The arrows indicate the direction of medium flow in the pipe.
Preferred mounting angle
Optimum mounting angle
Acceptable mounting angle
Mounting angle to be avoided
Forbidden mounting angle
Endress+Hauser
OUSAF22
Environment
Ambient temperature range
0 to 55 C (32 to 130 F)
Storage temperature
Humidity
5 to 95 %
Degree of protection
Process
Process temperature
Process pressure
Max. 100 bar (1450 psi) absolute, depending on the material, pipe size and process connection of
the flow assembly
Mechanical construction
Design, dimensions
B
A0028304
7
A
B
C
Sensor module
Dimension of lamp, depends on lamp type, see table
Dimension of detector, see table
Assembly, see Technical Information for assembly
Lamp type
Dimension A in mm (inch)
33.78 (1.33)
Gas-filled lamp
33.78 (1.33)
Detector type
Dimension B in mm (inch)
101.6 (4.0)
The total length of the sensor module is derived from the lengths of the lamp, the detector and
the assembly.
The dimensions of the OUA260 assembly are provided in Technical Information, TI00418C.
When connecting the sensor cable, make sure you maintain an additional distance of 5 cm (2")
on both the lamp side and the detector side of the sensor.
Endress+Hauser
OUSAF22
Weight
Materials
Sensor housing
OUA260 assembly
Detector
Filter
Declaration of Conformity
The product meets the requirements of the harmonized European standards. As such, it complies
with the legal specifications of the EC directives. The manufacturer confirms successful testing of the
product by affixing to it the mark.
Ex approvals
ATEX II 2G Ex db IIC T5 Gb
FM Cl.1, Div. 1, Groups B, C, D
FDA conformity
All non-metal parts in contact with medium, such as rubber and plastic parts, meet the requirements
of FDA 21 CFR 177.2600. The plastic and elastomer parts of the sensor in contact with medium
have passed the biological reactivity tests according to USP <87> and <88> Class VI.
Ordering information
Product page
www.endress.com/ousaf22
Product Configurator
On the product page there is a "Configuration" button to the right of the product image.
1.
2.
Select all the options to configure the device in line with your requirements.
In this way, you receive a valid and complete order code for the device.
3.
Export the order code as a PDF or Excel file. To do so, click the appropriate button on the right
above the selection window.
For many products you also have the option of downloading CAD or 2D drawings of the
selected product version. To do so, click the "CAD" tab and select the desired file type using dropdown lists.
Scope of delivery
The scope of delivery consists of the following depending on the version ordered:
sensor
Detector and lamp module without flow assembly or
Detector and lamp module mounted on OUA260 flow assembly
Operating Instructions
Endress+Hauser
OUSAF22
If the sensor is ordered with a transmitter, the complete measuring system is factory-calibrated and
shipped as one package.
If you have any questions, please contact your supplier or your local sales center.
Accessories
The following are the most important accessories available at the time this documentation was
issued. For accessories not listed here, please contact your service or sales office.
Flow assembly
OUA260
Flow assembly for hygienic sensors
For sensor installation in pipes
Materials: stainless steel 316, 316L or Kynar (other materials available on request)
Wide variety of process connections and path lengths available
Product Configurator on the product page: www.endress.com/oua260
Technical Information TI00418C
Cable
Endress+Hauser
www.addresses.endress.com