Measuring Coaxial Cable Impedance Using Oscilloscope To Analyze Waveform
Measuring Coaxial Cable Impedance Using Oscilloscope To Analyze Waveform
Measuring Coaxial Cable Impedance Using Oscilloscope To Analyze Waveform
Page
(1) ABSTRACT 2
Kicker Magnets 3
Transmission Lines 6
(3) THEORY 6
Transmission Line as Pulsed Power Modulator 6
Transmission Line Equations 8
High Frequency in Coaxial Cables with Losses 10
Losses in Pulses through Transmission Line 11
Attenuation of Waveform due to Skin Effect 13
(4) MEASUREMENT 14
Experimental Apparatus 14
Measurement Method 15
(6) CONCLUSION 21
(7) ACKNOWLEDGEMENTS 22
(8) REFERENCES 23
1
Measuring Coaxial Cable Impedance using Oscilloscope to
Analyze Waveform
Tolani E. Owusu
ABSTRACT
The increasing use of electrical pulses in the transmission of power and data by cable has
resulted in a need for a better understanding of the electrical characteristics of a cable. Today,
the study of transmission line cable is no longer limited to idealized lengths of lines present but
also their properties as a function of materials and geometry. The finite speed of transmission
pulses and attenuation through coaxial cables plays an important role in the fast injection systems
distortion of arbitrary pulses through coaxial transmission line becomes a problem, especially
during power transmission from the storage line pulse generators to the fast kicker magnet that
deflects the particle beam. To better understand the cable properties, the pulse response of
coaxial cable line was measured using an oscilloscope. Then analysis of different waveforms
was undertaken.
*
Operated by the Universities Research Association under contract with the U.S. Department of
Energy.
2
First, attenuation and kick strength rise time for transmission line kicker is described.
Next, the theory of conductor and dielectric losses in high voltage pulse cable is reviewed.
Finally, the decrease in amplitude and deformation of waveform due to skin effect is examined.
The analysis of waveform shows that pulses traveling in lossless line pulse in traverses-
electromagnetic mode (TEM) without attenuation and distortions are only mathematical
abstractions. However, all physically realizable cables suffer from attenuation and distortion of
pulses, with skin effect in the conductors being the predominate losses.
through coaxial cables is vital. Specifications of kicker magnets for particle accelerators
typically require fast rise times and good field uniformity. Kicker magnets are used to inject and
extract beams. As a case in point, 1MHz beam chopper is use to create gaps in the beam during
which the field in the kicker magnet must rise or fall from 1% to 99% in order to minimize beam
losses. The time interval of the gap created is approximately 50ns long.
Another example is an injection kicker for the Main Injector. It must have a rise time
(1% to 99%) of 50ns, a field uniformity of 1% over 1.2us and a fall time (99% to 1% of
150ns). In general, the required rise and fall time of each kicker magnet varies from accelerator
to accelerator. Some kicker magnets consist of LC cells with each cell sandwiched between high
voltage capacitance plates. The characteristic impedance for such kicker magnet is defined in
terms of the inductance and capacitance of each LC cells. The purpose of an extraction kicker
magnet is to fully deflect (kick) bunches of particles without disturbance to the preceding or
following bunches. On the other hand, an injection kicker does the inverse. For instance, it
places a particle beam from septum magnet onto an equilibrium trajectory. Again, kicker
3
magnets are fast pulsed magnets and are pulsed from a high voltage pulse generator. All of the
injection and extraction kicker magnets at Fermilab are designed to have characteristic
reflections in the system, the characteristic impedance of all the components are matched closely.
The output cable from the kicker magnet shown in figure 1 is connected to main switch
thyratron, which is connected to a pulse-forming network. In the thyratron switch off state, the
magnet is not connected (see figure 1) and as a result, it is possible to change the pulse forming
network slowly to a high voltage (kV). The thyratron switch is then closed and a large current is
4
As a fact, the electrical characteristics of the coaxial cables used in the accelerator at
Fermilab have to meet very rigid specifications. Hence, measuring accurately the properties of
capacitance, inductance and other characteristics are vital. In general, the rise time and
attenuation of transmitted pulses must satisfy certain minimum conditions in order to inject or
The majority of the circular accelerator is a bending magnetic field (see figure 2).
5
Figure 2. Bending Magnetic Field
The beams need to be steered between different accelerators. Due to economical reasons, strong
electric fields cannot be produced for high-energy accelerators. Also, to maintain high interaction
rate of experiments, it is essential to accurately steer these beams (see figure3). Therefore,
the distortion of beam diagnostics and monitoring signals needs to be known. Hence, the
theoretical methods, which allow computation of pulse response for different types of
coaxial cables response to an input signal with an arbitrary frequency spectrum is relatively
complex and needs extensive calculation [1]. However, the frequency content of pulses applied
6
is usually concentrated with a limited frequency range. As a result, it is sometimes possible to
find simple solutions that describe the behavior of cables with good approximation. Hence, an
analysis was undertaken to study frequency and pulses response of coaxial transmission cable
Transmission lines play a central role in extraction and injection of beams. For instance,
they are used to transmit pulses from the pulse forming network through the switch thyratron to
the kicker magnet (see figure 1). A transmission line is a distributed capacitor and inductor.
Energy is stored in the line when the center conductor is charged to high voltage. They have a
characteristic impedance that produces a constant voltage output pulse when discharged into a
resistive load, such as at the end of kicker magnet. The output pulse has amplitude of one half of
the charging voltage when the characteristic impedance has the same value as the resistance of
the load. Again, transmission lines provide one of the most essential exceptions to the low
frequency principles that the signal source should have a source impedance small compared with
the impedance of the load being driven and that the load should present input impedance large
compared with the source impedance driving it. The equivalent rule for transmission lines is that
the load and (the source) should present impedance equal to the characteristic impedance of the
II. THEORY
Most accelerator applications for pulse modulator line require a constant voltage pulse. A
critically damped waveform is the closest a modulator with a single capacitor and inductor can
7
approach constant voltage. Producing the desired pulse may be understood in terms of theory of
wave propagation.
Consider a transmission line with a current-fed source that produces a rectangular pulse.
At the instant when the switch is opened, the current IO from a constant current source divides
equally between the line and the resistance load if the resistance of the two are equal. A voltage
wave of amplitude IoZo traveling down the line is reflected with reversal of sign at the short-
(2.1)
2 =
when the reflected wave reaches the input end, the voltage drops immediately to zero because the
line is matched by the resistance RL=Zo and this results no reflection at the input end. However,
the line is fully charged with magnetic energy since a constant current of I o is flowing through it.
The voltage pulse generated at the load during the charging period has amplitude of
(2.2)
IoRL
and duration of 2Nevertheless, if the circuit between the constant current source and the
transmission line is broken, a voltage pulse is generated by the resulting discharge of the
magnetic energy stored in the line. The total energy stored on the line is LoIo where Lo is the
total distributed inductance of the line. In general, the energy is equal to the pulse energy
dissipated in the load. In contrast, when the transmission line used voltage-fed current pulse-
generating source, the far end of the line is open-circuited instead of short-circuited. In voltage-
fed line, the line is charged to a voltage Vo, which produces as current pulse on matched load of
amplitude Vo/2RL.
8
In summary, a discharge of a transmission line through a shorting switch such as the
thyratron into a matched resistive load produces a constant voltage pulse. The magnitude of the
(2.3)
l
t 2
v
Because the properties of coaxial cables such as inductance, capacitance, resistance and
individually. As a result, a brief review of transmission line theory is essential. The time it takes
a transmission line to respond to an input signal is calculated from transmission line equations
(2.4)
V
( R L ) I
z t
I
(G C )V
z t
Where
R = resistance/meter
C = capacitance/meter
L = inductance/meter
G = conductance/meter
9
It must be emphasized that in general, R is not the inverse of G. L, C, G, R, are per unit
For spectral amplitudes of voltage and current Vw, Iw, it follows from Eq 2.3 that
(2.5)
Vw( z )
( R jwL ).Iw( z ) Zs ( w).Iw( z )
z
Iw( z )
(G jwC ).Vw( z ) Yp ( w).Vw( z )
z
The above equations are valid if the series impedance/length Zs and the parallel
(2.6)
where Vw+ and Vw- are the voltage amplitudes of the waves moving in the +Z or Z direction
(2.7)
Zs ( w)
Zo
Yp ( w)
10
and the propagation velocity is
(2.8)
1
V
ZsYp
A transmission line, whatever its form, has characteristics impedance Zo; meaning a pulse
moving along a line has a ratio of voltage to current. From equations (2.4) and (2.5) we see if
(2.9)
L
Zo
C
where Zo is the characteristics impedance. For small losses in coaxial cables propagation
(2.10)
1 R G
( R jwL )(G jwC ) jw LC [1 ( )]
2 jwL jwC
or j
R 1
GZo
Zo 2
LC
where is the attenuation constant in the cable conductors and is the phase constant.
11
D. Losses in Pulses through Transmission Line
The basic structures of the two coaxial cables we analyzed are shown in figure 5. It is
outer cylinder Ro. The conductor is separated by a medium with dielectric constant and linear
12
(2.11)
2
C
Ro
ln( )
Ri
and
Ro
L ln( )
2 Ri
where
dielectric
of the insulator
permeability of the dielectric
By substituting equations 2.11 into 2.8 the velocity propagation in coaxial transmission line v is
(2.12)
From the above equation it is realized that the velocity in the medium is speed of light. For all
1 c
v
LC
TEM transmission lines the geometric factors in C and L cancel, therefore, the propagation
velocity is determined only to be the properties of the medium filling the line. Further analysis
also show that at any point on the transmission line voltage is linearly proportional to current,
given by
(2.13)
V = IZo
Attenuation is a measure of the loss of energy between the transmitting and receiving
terminal. The loss of energy depends on many factors such as frequency, conductor size and
spacing. Energy loss is high at higher frequencies and all higher frequencies are increasingly
13
attenuated. The reason is that at high frequencies changing magnetic fields does not penetrate
very deep into a good conducting material. As a result, the effective depth of the current
penetration in the conducting cable is negligibly small as compared to the cable dimensions.
Hence, the current density is non uniform a phenomenon known as skin effect. In transmission
lines, the amplitude of the wave attenuates as the wave propagates along the transmission line.
The distance through which the wave amplitude is reduced by a factor of 1/e (about 37%) is
known as skin depth or penetration depth of the line and is given by the equation
(2.15)
2
As the skin depth gets smaller with respect to the inner and outer conductors the attenuation A
(2.16)
However, for a matched coaxial cables analyzed, neglecting pure cable delay, the distortion of
1 1 1
A ( )
Ra Ri ln( Ra )
Ri
A
w
signal and cable rise-time is given by
(2.17)
o ( Al )
The analysis of equation 2.17 reveals two important ideas. First, rectangular pulses of
weight 1 with pulse length much smaller than to are severely attenuated. Second, rectangular
pulses with pulse length much larger than to can be approximated as a unit step input pulse with
14
(2.18)
1
Vout ( y ) erfc ( )
2 y
where
2
erf ( x ) 1 erfc ( x )
exp(-u)
where
t
y
o
For further information regarding equations 2.16, 2.17 and 2.18, the reader is referred to
reference 1.
III MEASUREMENTS
A. Experimental Apparatus
In order to plot the desired electrical signal against a linear time base, the most common
and widely applicable type of indicator on the screen of a Cathode Ray Tube (CRT), is the
oscilloscope. The oscilloscope makes it possible to study the pulse to pulse variations. The
digital oscilloscope LC 584 was used in our experiment. It differs from conventional
oscilloscopes principally, because it is particularly designed for the observation of fast pulses and
hence it utilizes fast sweeps that are synchronous with the signal to be observed. Again, several
adjustments are incorporated for convenience such as a means of turning off the filtering in order
to see the actual signal, varying the delay of the sweep and the signal and amplitude. This
signal as a function of time. On the screen of a cathode ray tube (CRT) is a plot of electrical
signal against a linear time base. The digitizing oscilloscope works by discretizing the signal,
storing it in digital memory and then reconstructing it on fluorescent screen. The data can also be
15
B. Measurement Method
In our experiment, pulse responses to several coaxial cable lines were studied; however,
we only concentrated on the AA5966 cable in MI-52 and the AA8740 spool of special cable in F-
17. Not in any particular order, we started our analysis with the AA5966 kicker cable. First,
reading the time domain reflectometer (TDR), the length of the cable was calculated to be
approximately 743ft (226.47 meters). Next, the digital low resistance Ohmmeter (DLRO)
measured the DC resistance of the center and outer conductor to be 0.112 Ohms and 0.194 Ohms
. For the observation or measurement of the pulse output from the pulsar, it was necessary
to find a means by which the voltage pulse could be applied as a signal to the CRT. In respect to
this, a resistive divider was introduced. The resistive divider was a combination of a number of
resistance in series and parallel. To obtain the oscilloscopic presentations of the voltage pulse,
two different trials were performed using the same coaxial cable. The reason, to reduce the
In the first trial A, a resistive divider output port 1 was connected to the oscilloscope
while the other output port 2 was connected to the input of the cable under test. The output of
the AA5966 the cable under test was connected to channel 2 of the scope. With the second trial
B, using the same cable, the output of port 2 of the divider was connected to the scope while the
16
Figure 5. Circuit diagram of Cable under Test
output of port 1 of the divider was connected to the input of the cable under test. Again, the
output of the cable under test was connected to the scope. These two trials were performed in
order to measure the time and amplitude of the cable relative to the input voltage. The output
signals of the cable under test were measured and stored in the oscilloscope. Figure 7 shows the
average calculations of experimental input and output amplitudes with respect to time and the
Similarly, the above procedure was repeated for the AA-8740 spool of special cable at F-17.
The measured cable length was approximately 793ft (225.25 meters). The DC resistance of the
center and outer conductors were measured to be 0.222 Ohms and 0.230 Ohms respectively. Again,
17
Figure 7. Coaxial Cable AA5966
18
Figure 8. Coaxial Cable AA8740
Figure 9. Comparison of AA5966 (smaller cable in green) and AA8740 (larger cable in red)
Pulse is a departure of some electrical quantity such as voltage, current or power from
zero or equilibrium value. In general, the number of pulse shapes actually encountered in
practical circuit is legion. However, the irregularities in pulse shapes that are encountered make
it apparent that determining pulse voltage is limited to special applications where the shape is
known and the nature of the pulse response to that particular pulse shape being measured is also
known. Since the measurement of coaxial cable impedance is sensitive to pulse shape,
oscilloscopic measurements are one of the best ways to interpret pulse response. In view of this
19
fact, the LC584 oscilloscope we used in our experiment provided flexibility for average
amplitude measurements and qualitative information regarding the shape of the pulse.
In a circuit where the input voltage pulse is of an irregular shape similar to that sketched
in figures 7 and 8, how does one measure the input voltage? In this experiment, we corrected
such equivalent pulse and approximated it as a square wave. The correction was done by
determining the input pulse amplitude at a point near the center top of the pulse. It may be
observed that the two cable attenuation is not constant with frequency and that there are some
distortions of pulses in both output average pulses. The distortions are particularly noticeable in
the rounding of the leading top corner of the pulse and slowing down at rate of rise as compared
to the calculated. Again, there is a decrease in the amplitude of the output pulses. In order to
find how much the pulse had decreased, we verified the theory for both calculated and
experimental pulse shapes. The calculated pulse shape was found by using the error function in
Several calculations were done using high attenuation of 8E-8Np/m and low 6.5E-
8Np/m. However, the best attenuation was 7E-8Np/m for the smaller cable AA5966. Not
surprisingly, as it was noted, the larger cable (AA8740) had the best attenuation of 5E-8Np/m
while its high and low were 5E-8Np/m and 4E-8Np/m respectively. Another striking thing that
was observed in AA8740 was the ripple at the top center of the experimental pulse. This could
2.16) from geometry was also calculated. The calculation for AA5966 was 9E-8Np/m with an
error of 15%. And the calculations for AA8740 was 6.2E-8Np/m with an error of 15%.
When comparing these two figures, note that both calculated and experimental pulse
shapes were obtained by using average values. The significant difference between cable
AA5699 and AA8740 is that the correspondence between the calculated and experimental pulse
20
shapes is closer in AA8740 than in AA5699. This difference occurred because larger cables have
fewer losses than smaller cables due to skin effect. Summarizing the discussion on pulse
forming line, it may be said that all physical system have unavoidable losses and distributed
capacitances and inductances with skin effect in the conductors being the predominate loss.
IV CONCLUSION
High losses of attenuation and distortion of pulse shape are major concerns of
transmission of power by the coaxial cable. The cable is a power supply to the kicker and high
loss pulses means the rise and fall time of the cable is too long which could cause a particle beam
to be kicked unintentionally. During extraction and injection of beam, timing and distance are
vital therefore; longer time causes fewer beams to be loaded in longer intervals.
Our analysis indicated that for perfect pulses such as square wave, bunches of beams
would be kicked in full trajectory without disturbance to the preceeding or following bunches.
Second, the development of pulse-forming line using smaller cable relative to larger cable is
economical but has higher losses due to skin effect. As may be anticipated, no network having a
finite number of elements can exactly stimulate a transmission line. However, we found that
larger coaxial cables such as the AA8740 have lower losses, and can store more voltages and
currents that could result more kick and less unwanted deflection of the beam from one end to
another.
Since larger cables have fewer losses and can store more voltages, our analysis indicates
that the geometry properties of coaxial cables must be investigated with care to determine how
closely the loss could be approximated in order to design the most useful and economical
transmission line.
21
ACKNOWLEDGEMENTS
This has been a great learning experience for me this summer. However, it has not have
been possible had it not be the immersed help, excellent discussions and guidance of my superb
supervisor, Chris C. Jensen. I would also like to thank my alternating supervisor Darren Qunell
for his discussions along the way. My thanks also go to Robert Hively and all the technicians in
the kicker group especially Ken Kellogg for our numerous trips to the data sites.
Finally, my sincere thanks to the SIST committee for selecting me to participate in this
program.
22
REFERENCES
[1] H. Riege, High-Frequency and Pulse Response of Coaxial Transmission Cables, with
conductor, Dielectric and Semiconductor Losses, CERN 70-4. Proton Synchrotron Department,
February, 4 1970.
[2] F. Kirsten and R. Wallace, Pulse Response of Coaxial Cables Lawrence Radiation Laboratory,
[3] Dworsky, Lawrence, Modern Transmission Line Theory and Applications. New York: J,
Wiley, 1979.
[4] Robert Grover Brown, Robert A Sharpe, William Lewis Hughes and Robert E. Post, Lines,
Waves And Antennas, The Transmission of Electric energy New York; J, Wiley 1961, 1973.
[5] Paul Horowitz and Winfield Hill, The Art of Electronics Cambridge University Press 1980.
23
24