C7510 01 2276
C7510 01 2276
C7510 01 2276
Reagent Set
Quality Control
Serum controls with known normal and elevated values should be run
routinely to monitor the validity of the reaction. These controls should be run
at least with every working shift in which Cholesterol assays are performed.
It is recommended that each laboratory establish their own frequency of
control determination. Quality control requirements should be performed in
conformance with local, state, and/or Federal regulations or accreditation
requirements.
Expected Values11
Recommended Range:
Desirable Cholesterol: <200mg/dl
Borderline-High Cholesterol: 200-239mg/dl
High Cholesterol: >240mg/dl
Performance
1. Linearity: 500 mg/dl
2. Comparison: A comparison between this procedure and a similar dry
powder reagent produced a regression equation of y = 1.02x + 0.8 with
a correlation coefficient of 0.999.
3. Precision:
Use by (YYYY-MM-DD) Lot and batch code Catalog number Manufacturer In vitro diagnostic medical device
Temperature limitation Consult instructions for use CE mark Authorized representative in the European Community