At 97553
At 97553
At 97553
|
TECHNICAL SUPPORT AND ORDERS
Tel: (33) 03 23 25 15 50
IVD IN VITRO DIAGNOSTIC USE
Fax: (33) 03 23 256 256
Reagent R1 1 mL
EXPECTED VALUES (2)
Working Reagent 1 mL
Direct Bilirubin mg/dL [mol/L]
Specimen or calibrator 20 L 20 L
Adult (and children over 5 days) < 0.2 [< 3.4]
Mix and read absorbance at 550 nm (540-560) after 1 minute at 37 C
Each laboratory should establish its own normal ranges for the (2 minutes maximum) against blank.
population that it serves.
Notes:
1. After 2 minutes, indirect bilirubin reacts slowly with diazotised
LINEARITY dichloroanilin and leads to over-estimated values.
Procedure n1: up to 20 mg/dL (342 mol/L). 2. To take into account the coloration of the working reagent, one
should perform a reagent blank.
Above, do not dilute the specimen: perform procedure n2. 3. Specific procedures are available upon request for automated
Procedure n2: up to 100 mg/dL (1710 mol/L) instruments. Please contact BIOLABO technical support.
Paediatric specimen: perform procedure n2
CALCULATION
Calculate the result as follows:
PERFORMANCES CHARACTERISTICS (PROCEDURE N1)
Within run: Medium level High level Between run Medium level High level With a calibrator (Procedure n1 only):
N = 20 N = 20
Abs (Assay - Blank) specimen
Mean mg/dL 1.13 2.54 Mean mg/dL 0.74 1.75 Result = x calibrator concentration
Abs (Assay - Blank) calibrator
S.D. mg/dL: 0.019 0.029 S.D.mg/dL: 0.017 0.032
C.V. %: 1.68 1.15 C.V. %: 2.3 1.84 with factor:
Detection limit: approximately 0.1 mg/dL Procedure n1: mg/dL = [Abs. Assay Abs. Blank] x 14.8*
mol/L = [Abs. Assay Abs. Blank] x 253*
Sensitivity for 1 mg/dL: 70 mAbs at 550 nm.
Procedure n2: mg/dL = [Abs. Assay Abs. Blank]] x 68.6*
Comparison studies with commercially available reagent:
mol /L =[[Abs. Assay Abs. Blank]] x 1173*
y = 1.0084x 0.527 r = 0.9926
*These factors should be used as a guide only and may vary with
instrument and the batch of reagent used. It is recommended to verify
it with elevated control serum.
(Procedure n1: use BIOLABO EXATROL-P, Procedure n2: use
BIOLABO PAEDIATRIC CONTROL).
REFERENCES
rd
(1) TIETZ N.W. Text book of clinical chemistry, 3 Ed. C.A. Burtis, E.R.
Ashwood, W.B. Saunders (1999) p. 1133-1137.
(2) Clinical Guide to Laboratory Test, 4th Ed., N.W. TIETZ (2006) p. 172-177
(3) YOUNG D.S., Effect of Drugs on Clinical laboratory Tests, 4th Ed. (1995)
p.3-90 to 3-110
(4) RAND R. N. and DI PASQUA A. A new diazo method for determination of
bilirubin. Clin. Chem. (1962) 8, n6, p. 570-578
(5) GOLUB M.: An automated method for determination of serum bilirubin.
Clin. Chem. (1964) 10, p. 399-405
(6) Henry RJ, Clin Chem: Principles and technics. Harper and Row.
p.592(1965)
(7) SRM: Standard Reference Material
Manufacturer Use by In vitro diagnostic Temperature limitation Catalogue number See insert Batch number Store away from light sufficient for dilute with