TRAX Manual en
TRAX Manual en
TRAX Manual en
Users Manual
WWW.MEGGER.COM
Users Manual
Contents
8 Update TRAX
.............................................................. 52
8.1 Upgrading..................................................52
Upgrading via Internet............................................... 52
Upgrading via USB..................................................... 52
Getting a TRAX (USB) upgrade via PC......................... 52
9 Specifications
.............................................................. 54
Specifications TRAX..........................................54
Index........................................................ 56
1 Introduction
Receiving Instructions
Check the equipment received against the packing
list to ensure that all materials are present. Notify
Megger of any shortage.
Examine the instrument for damage occurred in
transit. If damage is identified, file a claim with the
carrier at once and notify Megger giving a detailed
description of the damage.
The instrument has been thoroughly tested,
calibrated and inspected to meet rigid specifications
before being shipped. It is ready for use when set
up as indicated in this user manual.
Warranty repair
Equipment returned to the factory for repair must be
shipped prepaid and insured.
Contact your Megger representative for instructions
and a return authorization (RA) number.
Please indicate all pertinent information, including
problem symptoms.
Also specify the serial number and the catalog
number of the unit.
2 Safety
2.1 General
Caution
Important Caution, as used in this manual, is defined as
Read and comply with the following in- a condition or practice which could result in
structions. damage to or destruction of the equipment or
Always comply with local safety regula- apparatus under test.
tions.
3. The power input plug must be inserted only 2.2 Safety instructions
into a mating receptacle with a ground contact.
Do not bypass the grounding connection. Any 1. It is not possible to eliminate all potential
interruption of the grounding connection can hazards from, and in using, electrical test
create an electric shock hazard. Determine that equipment. For this reason, every effort has
the receptacle is properly wired before inserting been made to point out in this instruction
the plug. manual the proper procedures and precautions
to be followed by the user in operating this
equipment and to mark the equipment itself
with precautionary warnings where appropriate.
It is not possible to foresee every hazard which
may occur in the various applications of this
equipment. It is therefore essential that the user,
in addition to following the safety rules in this
manual, also carefully consider all safety aspects
of the test before proceeding.
2. The test set and the specimen to which it is
connected are a possible source of high-voltage
electrical energy and all persons making or
assisting in tests must use all safety practice
precautions to prevent contact with energized
parts of the test equipment and related circuits.
3. Persons actually engaged in the test must stand
clear of all parts of the complete high-voltage
circuit, including all connections, unless the
test set is de-energized and all parts of the
test circuit are grounded. Persons not directly
involved with the work must be kept away from
test activities by suitable barriers, barricades, or
warnings.
4. Treat all terminals of high-voltage power
equipment as a potential electric shock hazard.
There is always the potential of voltages being
induced at these terminals because of proximity
to energized high-voltage lines or equipment.
5. Always ground connection points of the test
specimen before connecting any leads from the
test set. Whenever possible, always keep one
side of the test specimen grounded at all times.
Always use a safety ground stick to ground any
high-voltage conductor.
6. The ground connection on the test set must
be the first made and the last removed. Any
interruption of the grounding connection can
create an electric shock hazard.
7. Make sure that the instrument is properly
grounded, both through its AC power cord and
through the ground connector. The AC power
cord is the disconnecting device.
8. Always disconnect test leads from test specimen
before attempting to disconnect them at the
test set.
9. High-voltage discharges and other sources of 3. Read and understand Safety in the User Manual
strong electric or magnetic fields may interfere before performing any service.
with the proper functioning of heart pacemak- 4. Routine maintenance is all that is required for
ers. Persons with heart pacemakers should these test sets. The cables and connector panel
obtain expert advice on the possible risks before should be inspected frequently to be sure all
operating this equipment or being close to the connections are tight and all ground connec-
equipment during operation. tions intact.
10. All persons making or assisting in tests must 5. The appearance of the test set can be main-
use all practical safety precautions to prevent tained by occasional cleaning of the case,
contact with energized parts of the test equip- panel and cable assemblies. The outside of the
ment and related circuits. Also follow all local carrying case can be cleaned with detergent and
and company safety requirements. Persons water. Dry with a clean, dry cloth. The control
actually engaged in the test must stand clear of panel can be cleaned with a cloth dampened
all parts of the complete high- voltage circuit, with detergent and water. Do not allow water
including all connections, unless the test set is to penetrate panel holes, because damage to
de-energized and all parts of the test circuit are components on the underside may occur. A
grounded. Persons not directly involved with the household all-purpose spray cleaner can be used
work must be kept away from test activities by to clean the panel. Polish with a soft, dry cloth,
suitable barriers, barricades, or warnings. taking care not to scratch the display screen
11. Safety is the responsibility of the user. cover. The cables and mating panel receptacles
12. Misuse of this high-voltage equipment can be can be cleaned with isopropyl or denatured
extremely dangerous. alcohol applied with a clean cloth.
13. The purpose of this equipment is limited to use
as described in this manual. Do not use the
equipment or its accessories with any device
other than specifically described.
14. Before making any connections, make sure that
the instrument is de-energized and that all parts
of the test circuit are properly grounded.
15. Never connect more than one output at the
time. All outputs are energized by the same
amplifier and therefore all outputs are energized
simultaneously.
16. Operation is prohibited in rain or snow.
17. Do not use the test set in an explosive
atmosphere.
18. A qualified operator should be in attendance
at all times while the test equipment is in
operation.
19. Observe all safety warnings marked on the
equipment.
20. Corrective maintenance must only be performed
by qualified personnel who are familiar with the
design and operation of the test set and the
hazards involved.
Maintenance
1. DISCONNECT the MAINS plug before any
cleaning or maintenance.
2. Maintenance should be performed only by
qualified personnel familiar with the hazards
involved with high-voltage test equipment.
7
1
8
2
3
4
9
5
10
11
12
6
13
1 2
3 9
10
4 11
5 12
3
1 2 3 4
1. TRANS 5
General input for analog transducers and low
level analog signals e.g. motion transducers,
Rogowski coils etc.
2. CONTROL
close/open contacts for OLTC and circuit-breaker 4
control (up-down, close-trip)
3. TRIG IN
External trig input for starting measurements or
recording based on external event.
1. VOLTAGE INPUTS
4. TIMING Four channels 0-250 V AC, 0-350 V DC
Binary inputs for timing measurements in timer
2. CURRENT INPUTS
and relay testing applications when used as a
Four channels 0-10 A AC, 0-10 A DC
timer. A and B inputs dedicated for Start and
Stop. Note Voltage and current cannot be measured on
the same channel simultaneously.
3. FUSES
4 x 500 mA/25 V AC fast, can be changed from
the outside, protects the low current gain shunt.
Inside the top panel, there are 4 x 15 A/250 V AC
fast fuses for the high current gain step
4. DC INPUTS
These two channels (R1 & R2) are designed for
measuring low DC voltage when measuring con-
tact and winding resistance, using the 100 A or
1 to 16 A DC current outputs. If the channels are
used for measuring AC, max input is 40 V RMS.
5. LED INDICATORS
Red LEDs indicate which channel to connect to
depending on what App is used.
4.1 General how many taps. From that the TRAX app is mastering
the test and you have to perform the test following
TRAX measurements are collected together in tests the connections described by the unit to get the
and test sessions. A test is often containing several automatic assessment of the test.
individual measurements and a TRAX test ses-
Also here the app measures one setup at a time.
sion/file/report often contains several tests performed
As an example, if the test object is a three-winding
with different apps.
transformer, TTR measurements may be performed for
three setups, HV-LV, HV-tertiary and LV-tertiary. Each
Manual and configured tests of these is done as a separate test with a separate test
table, collected together in the same test session.
Manual Control
Manual tests can be done by using Manual Control Test object information
where you define exactly what generator to use,
(nameplate)
type of test signal, how results are measured and
how parameters are calculated. This gives unlimited Test object/nameplate information must be entered
possibilities to perform almost any AC or DC test on for any configured test and will be part of the test
any electrical component within the limits of TRAX report. The mandatory information needed for a
generating capabilities. certain app is asked for and added to the report. The
next app will have the same information automatically
Results from Manual Control measurements for a
but may need some additional information to be
certain setup are collected in a result table. If the test
added. After all tests you can go to the report and
setup is modified due to change of measurement
add any information that is still missing.
channels and/or calculated parameters, the new
results are collected in a new test/results table.
Create a test session in advance
Manual test If a test session needs to be defined before the test is
It is also possible to use TRAX as a manual standard actually performed, the workflow is the following:
instrument. This operation mode is defined as 1] Open the first app and define the test object,
Manual Test/No configuration and is available define parameters such as: transformer vec-
in all the transformer testing apps. Compared to tor group.
Manual Control, this operation mode is locked to a
certain application. As an example, winding resistance 2] Name and save (the empty) test.
measurements in manual mode is limited to using any 3] Go to the report and fill in all necessary in-
of the three DC current generators and one or two formation for the test object and all planned
DC measurement inputs. tests.
Results from Manual Test measurements for a certain 4] Open the next app to create a test table
setup are collected in a results table. If the test setup
is changed by e.g. changing from single to dual
5] Save and go to the next app. Continue as
above for all desired apps.
channels (simultaneous winding magnetization)
winding resistance, the new results are collected in a 6] Finally check the report once again to see
new test/result table. that all the needed data is entered and that
all planned tests are there.
Configured test 7] Confirm/save and close.
Configured tests are defined by entering information
about the test object, as is the case of vector group
and configuration of a power transformer, if there is
a tap-changer or not and if so on what winding and
At the site
1] Load the report that was created for the test
object.
2] Go from report to a certain app to perform
the specific measurements as guided by
TRAX.
3] Save and go to the next app. Continue as
above for all desired apps.
5 Operating instructions
About
Home screen
Report view
Selecting app. If this is the first app
selected after starting the unit, TRAX For selecting apps to be shown under
prepares itself for a new test session. corresponding menu tab.
App If previous tests are performed, TRAX
continues with the same session (see Safe eject of all USB drives connected to the
section Data handling and reporting) TRAX
New test file/session/report
Turn off TRAX
New 1] Press for 3 seconds to turn off TRAX.
Load a saved test session/report
Global settings
Load 1] Press for the global settings.
Load a saved test as template for a new 2] Press the button + GUI Settings
test session A window with the following fields opens.
Template
Global settings for all apps
Global
settings
Updates
TRAX stand-alone, direct connect
Release TRAX main SW version
track
Download Get the available file
Install Selected update is installed
Updates Apps
TRAX stand-alone via USB 1] Press to access the different apps.
Insert a USB stick with a valid TRAX update (from remote
update)
Source Select USB
Download Get the available file from USB
Install Selected update is installed
Network
Settings for TRAX remote operation. Not available
in offline mode.
Report
Test Define naming for the first ID fields in
My TRAX In My TRAXyou can place
Object the report.
your favourite apps for
Header Select fields to be used for identifica-
easy access by opening the
Fields tion in TRAX.
Tools screen and activate a
certain app.
Safety Power Transformer Apps suitable for testing
power transformers.
Interlock Interlock 2 can be selected as manda-
Current Transformer Apps suitable for testing
tory for all TRAX generators/outputs. If
current transformers.
turned off, interlock 2 is only mandato-
Voltage transformer Apps suitable for testing
ry for 2 kV and 12 kV* generator.
voltage transformers.
*TDX120 optional accessory
Substation Apps suitable for general
Ground Can be turned on/off for all generators testing in substations.
loop except 2 kV
detector
Note: Short description
Settings are not available in offline mode
The winding resistance app is used for
measuring DC resistance in transform-
Hardware Configuration er windings and other objects with
Winding high inductance.
Internal Transformer size
resistance
Turns ratio measurements on trans-
About former windings (TTR).
About HW and SW version information
Turn ratio
Temperature Temperatures in TRAX internal parts
Contact resistance app is used for
Legal Legal information
DC low resistance measurements on
Notices
resistive loads.
Contact
resistance
Excitation current app is used for
measuring current and impedance on
one side of a transformer with the
Excitation opposite windings open.
current
Home
Report
Return
Apps
Generator settings
Warning
The Manual Control is not designed for
resistance measurements on inductive
loads.
App settings
1] Press
Example:
Channels Active current clamp, labeled 10 mV/A
Transducer Int
Transducers can be used to convert a measurement
signal to an appropriate current or voltage signal to be
measured by TRAX, e.g. active current clamps (current
to voltage).
Start/Stop measurements
1] Press to start the measurement.
Output signals and selected measurement signals are
continuously displayed on the analog meters and in
the results fields.
Oscilloscope
1] Press
The oscilloscope can be used to monitor measurement
signals.
2] Press Inputs (red square) and select meas-
urement channels for display.
3] Press Freeze to hold and show the oscillo-
scope picture.
4] To close the oscilloscope, press
2] Select a DC current generator 1, 16 or 100A 10] Select an appropriate test current suitable
output (pending the resistance of the test for the test object.
object. Max resistance is about 10 k when Always use as high current as possible but
generating 5 mA from the 1 A generator). avoid unintentional heating of the test object.
7] Wait for results to stabilize or manually con- 4] Press and select e.g. Inductance, Imped-
trol the voltage and observe the excitation ance, Reactance or other parameters to get
current for e.g. determine saturation/knee- appropriate data for the test.
point. 5] Connect current leads and voltage sense
8] Press and read the results. leads to 0-10 A output and channel 1 voltage
input respectively. At sample, place voltage
Note The description above is using an internal cur- sense leads inside current connectors.
rent measurement channel and measures the
total current generated into the test object. 6] Short appropriate low voltage winding (e.g.
(GST-GND) first measurement of a YNyn0, measure 1U-
If the test object has two parallel parts e.g. a 1N, short 2U-2N).
winding in a delta configuration, the mea-
sured value is the current going through one 7] Select an appropriate test current for the test
winding in parallel with two series windings. object. For power transformer windings, typ-
For measuring excitation current on a single ical test current are 1 to 5 A for short-circuit
winding in a delta configuration, an external impedance/leakage reactance measurements.
current measurement can be used and by
Note Max compliance voltage is 250 V. When mea-
grounding correctly a UST measurement can
suring small transformers with high winding
be performed.
resistance you need to select a lower current,
(typically 100 mA or lower) to not trip the
Warning generator.
When using 2.2 kV output and external
8] Press
ammeter, its is ABSOLUTELY MANDATORY
to ground black terminal and connect the 9] Wait for results to stabilize.
external current measurement channel 10] Press and read results.
to the grounded side of the generator/
winding. 11] Move to next phase and continue the test.
Note This measurement can also be performed
by using the 250V generator and adjust the
voltage until the preferred test current is
achieved.
Zero-sequence impedance
measurements
1] Select 0-250V AC generator and 55 Hz.
2] Select Gen I measurement channel, set to
freq, to capture an hold the excitation cur-
rent and use the data for calculations.
3] Select Ch 1, set to freq, for voltage measure-
ment.
(Gen U can be also used for the measurement
4] Press . 4] Press
For calculation 1, select ratio / (division) For calculation 1, select ratio / (division)
between Gen U and Ch 1. between Gen I and Ch 1.
If phase deviation is to be recorded, for If phase deviation is to be recorded, for cal-
calculation 2, select Phase between Gen U culation 2, select Phase between Gen I and
and Ch 1. Ch 1.
5] Connect CT terminal S2/X2 to ground. 5] Connect CT terminal S2/X2 to ground.
Connect black Ch 1 terminal to terminal S2/
6] Connect black generator terminal to S2/X2
X2 and blue Ch 1 to terminal S1/X1.
and green (250V)/red (2200V) generator ter-
minal to CT terminal S1/X1. 6] IMPORTANT
Connect P1/H1 (or P2/H2) to ground.
7] IMPORTANT
Connect 200A/800A current output terminals
Connect P1/H1 (or P2/H2) to ground.
to primary winding, black terminal to P2/H2
Connect Ch1 terminals to primary winding,
and red terminal to P1/H1.
black terminal to P2/H2 and red terminal to
P1/H1. 7] Select an appropriate test current, usually
rated primary current or a fraction of rated
8] Select an appropriate test voltage, best accu-
primary current.
racy is achieved at about 75% of saturation
Make sure expected resulted secondary cur-
voltage.
rent is lower than 1 A (IEC) or 5 A (IEEE).
9] Press to start the generating.
8] Press to start the generator.
10] Wait for results to stabilize
9] Wait for results to stabilize
11] Press and read results.
10] Press and read results.
1] Select 0-200 A (or 0-800 A if TRAX 280) gen- Note The test mode is GST-GND which means that
erator. the total current to ground will be measured.
The test result will be affected by all stray
Select frequency, typically 50 or 60 Hz.
capacitors including e.g. cables and surface
2] Activate Int Gen I, configure it to Freq. current. To estimate the effect of cables (typ-
ically 50-100 pF), measure the cables without
3] Activate channel Ch 1, configure it to Current
connecting to the specimen.
(A) and Freq.
Important
Read and comply with the safety instruc-
tions 2 Safety on page 8.
Always comply with local safety regula-
tions.
1] Press
Temperature On/Off
correction Measured values will be automat-
ically corrected to the reference
temperature.
Object temp Enter test object temperature.
(C)
Reference Reference temperature to where the
temp (C) resistance will be adjusted to.
Material For copper, aluminum or as customer
coefficient defined.
Custom Only applicable when Material
coefficient coefficient is set to customer defined.
Settings
1] Press DualGround
Make desired settings for the following subjects,
Ramp, Temperature correction DualGround and
Integration.
Ramp
DualGround On/Off
Clamp-on ammeter Set the ratio for the clamp-on
(mV/A) ammeter (mV)
(optional accessory,
connected to R2
Ramp up time (s) 1, 2, 3, 5, 10 or 20 input)
Constant current time (s) 3, 5, 10, 20, 60 or 3600 Show current On/Off
Ramp down time (s) 0.3, 1, 2, 3, 5, 10 or 20
1] Press
+ / raise
- / lower
The winding resistance app is used for measuring
DC resistance in transformer windings and other
objects with high inductance. The app supports single capture/measure data without stopping the
channel and dual channel (Simultaneous Winding generator (for OLTC measurements).
Magnetization, SWM).
Activating + or - buttons closes the contact for about
Recommended test current ranges 500ms. After one operation the contacts are blocked
for about 2 seconds before the next operation is
1A 5 mA1 A Resistance range 1m10k
possible.
16A 1A16 A Resistance range 16050
100A 10 A100 A Resistance range 105 Run two wires from each output contact and connect
them in parallell with the manual operating contacts
Note When testing transformer windings, the test
current should be sufficient to saturate the for raise/lower tap (at the tap-changer control cabi-
core and thus minimizing inductance. This net). Max carrying current (short-term) is 35 A.
happens typically at about 1% of the rated
winding current. Avoid testing at >15% of
rated current since this may cause heating
which will affect the accuracy. Typical test
currents are 1-15% and using 5-15% will give
fast and stable readings.
CAUTION
The no power emergency discharge may
take significantly longer time due to
lower discharge voltage. Make sure to
wait for a sufficient time (>2 minutes for
a large transformer) before removing any
cables.
HV measurements Settings
1] Press
LV measurements
Test Parameters
2] Select test parameters and make settings.
WARNING
Do not remove any cables until discharge
is finalized.
1] Press
Step Vs reduction per demag cycle
(default 50%)
Display % remanence or absolute values in Vs
(default %)
Step-by-step instructions
1] Connect generator cable and R1 measure-
ment cable to the transformer HV terminals.
If YN, connect to H2/V to neutral, if Y or D
connect H1/U to H2/V.
2] Select generator.
3] Select demagnetization current, normally the
same as the test current, about 1% of rated
winding current. TRAX turns ratio app determines the transformer
turns-ratio as defined by international standards. The
4] Press to start test app provides an excitation test voltage to the primary
5] Demagnetization starts and stops automati- windings of the transformer and measures simulta-
cally when finished. neously the voltage at the corresponding secondary
winding. The voltage ratio is displayed and compared
with the expected nameplate ratio.
TRAX measures ratio, phase deviation and excitation
current at the same time. Testing can be done at
power frequency, or preferably at a frequency differ-
ent from the power frequency to avoid interference.
Default test frequency is 55 Hz.
Transformer configuration
Transformer configuration and vector group is select-
ed by entering configuration via. keyboard or selecting
in the matrix.
If no configuration is entered the test will automati-
cally be defined as a manual TTR test on a 2-winding
transformer.
Step-by-step instructions
No configuration
1] Connect cables.
2] Select test voltage.
3] Press to start test.
4] Perform next measurement.
5] Save results.
Step-by-step instructions
No configuration
1] Select what generator to use, 250 V AC or
2200 V AC.
2] Connect the generator leads to the HV wind-
ing.
3] Press to start the first test.
4] Readings will be displayed automatically
when stable and generator will stop.
5] Continue with next test. New tests will be
added to the table. It is possible to re-meas-
ure a line in the table by activating the actual
row and start a new measurement.
6] When the test is finished the results can be
saved to a file/report.
Important
Read and comply with the safety instruc-
tions 2 Safety on page 8.
Always comply with local safety regula-
tions.
1] Press
Save
Export as xml
Export as text
Deleting file.
Can only be used if the file folder is
unlocked
Create new folder
Delete files
In the Load file window you can also select files to
delete.
1] Press to enable deleting.
2] Select on ore more files to delete.
3] Press
Logfiles can be downloaded to USB or PC with the upper right button. Filename is *.log. Format is tab delimited
text and the files can be directly imported as data from text in Excel. See example below.
Time Winding Tap Tap Voltage U (Fq) TTR Measured TTR % error I exc (Fq) Phase (Deg) Phase (Minutes)
16:01:04 H1-H3 / X0-X3 1 288,8 79,82298404 8,745039742 8,731460191 0,155282892 0,010902529 0,221320916 13,27925495
16:02:49 H1-H3 / X0-X3 1 288,8 79,73316426 8,745039742 8,731449598 0,155404028 0,010857631 0,218763181 13,12579086
16:03:06 H1-H3 / X0-X3 2 281,9125 79,79627549 8,536482051 8,520539455 0,186758382 0,011168756 0,218539378 13,11236269
16:03:26 H1-H3 / X0-X3 3 275,025 79,79326395 8,327924359 8,314507258 0,161109794 0,011634864 0,217935571 13,07613426
16:03:43 H1-H3 / X0-X3 4 268,1375 79,79542276 8,119366668 8,108744178 0,130829053 0,012166128 0,221951353 13,31708117
16:03:59 H1-H3 / X0-X3 5 261,25 79,77780771 7,910808977 7,897740809 0,165193822 0,012727199 0,225619454 13,53716726
16:04:23 H2-H1 / X0-X1 5 261,25 79,78636983 7,910808977 7,89808453 0,160848871 0,012395819 0,218229426 13,09376555
16:04:35 H2-H1 / X0-X1 4 268,1375 79,78636983 8,119366668 7,89808453 2,725362054 0,010952381 0,218229426 13,09376555
16:04:41 H2-H1 / X0-X1 4 268,1375 79,78636983 8,119366668 7,89808453 2,725362054 0,010952381 0,218229426 13,09376555
16:04:48 H2-H1 / X0-X1 4 268,1375 79,78636983 8,119366668 7,89808453 2,725362054 0,010952381 0,218229426 13,09376555
16:04:58 H2-H1 / X0-X1 4 268,1375 79,81087537 8,119366668 8,108414203 0,134893092 0,012499431 0,234196821 14,05180929
16:05:13 H2-H1 / X0-X1 4 268,1375 79,82208271 8,119366668 8,108918592 0,128680924 0,011717519 0,208185549 12,49113297
16:05:30 H2-H1 / X0-X1 3 275,025 79,79962282 8,327924359 8,314860947 0,156862758 0,011205049 0,204101436 12,24608618
16:05:46 H2-H1 / X0-X1 2 281,9125 79,80317883 8,536482051 8,520615935 0,185862464 0,010733944 0,203901405 12,2340843
16:06:03 H2-H1 / X0-X1 1 288,8 79,77118962 8,745039742 8,731616601 0,153494341 0,010298725 0,19919886 11,95193158
16:06:30 H3-H2 / X0-X2 1 288,8 79,81990417 8,745039742 8,723758047 0,243357324 0,007902342 0,154869073 9,292144371
16:06:38 H3-H2 / X0-X2 2 281,9125 79,81990417 8,536482051 8,723758047 2,193831082 0,007902342 0,154869073 9,292144371
16:07:01 H3-H2 / X0-X2 2 281,9125 79,83213248 8,536482051 8,513217914 0,272526038 0,008275373 0,161657851 9,699471065
16:07:18 H3-H2 / X0-X2 3 275,025 79,8004086 8,327924359 8,513275052 2,225652933 0,008245765 0,159178623 9,550717375
16:07:41 H3-H2 / X0-X2 3 275,025 79,7937044 8,327924359 8,307139828 0,249576371 0,008573294 0,157461986 9,447719172
16:08:06 H3-H2 / X0-X2 4 268,1375 79,81373307 8,119366668 8,101472628 0,220387141 0,008930984 0,159015948 9,540956886
16:08:31 H3-H2 / X0-X2 5 261,25 79,80348985 7,910808977 7,890445678 0,257411077 0,009323349 0,160623372 9,637402331
Note The logfiles are intended for emergency/backup use and are normally not used for reporting.
WiFi (optional) for running the instrument from an TRAX219 is only used with remote control.
external PC or other device.
Note For remote control, TRAX Control must be
Three USB ports for multi-purpose use: Transfer installed on the remote device.
data and reports to PC, use an external mouse or
keyboard, upgrading the unit from a USB memory The external device must have a Windows
stick etc. operating system version 7 or higher and
must have latest version Chrome web browser
installed.
Connecting to TRAX:
1] Connect the Ethernet cable to TRAX or to a
network where TRAX is connected.
2] Wait until your device has discovered an un-
known unit (normally by displaying a ? in
your network symbol)
3] Start TRAX Control.
4] After a while, the screen will display all TRAX
units found on the network.
5] Select the unit you want to connect to.
6] A dialog is started where you are asked to
confirm remote control by pressing and
holding the control knob on TRAX to allow
remote control.
8 Update TRAX
Important
Do NOT interrupt the updating process.
Important
Do NOT interrupt the updating process.
Important
The PC used to download the TRAX up-
date must have been connected to the
physical TRAX unit that is to be updated.
This connection must be the last connect-
ed TRAX. The procedure is mandatory for
identifying the correct update.
9 Specifications
Inputs
ANALOG
1234
Current 4 x 0-10A AC/DC
Voltage 4 x 250/350V AC/DC
R1 R2 2 x 0-50 V DC Intended for resistance
measurements but can be
used for AC voltage meas-
urement up to 40VRMS
TRANS Input for analog trans-
ducers and low level analog
signals
TRIG IN Contact or voltage sense
TIMING 3 x 0-10000 s Binary inputs for timing
measurements in timer and
relay testing applications. A
and B inputs dedicated for
Start and Stop.
Calculated / displayed parameters
Arithmetic +, -, *, /
Power P, VA, Q, S
Impedance R (DC), Z, Xp, Xs, Rs, Rp, Ls, Lp, Cs, Cp, phase
Derating at lower mains voltage
TRAX specification is valid at 230-240 V mains voltage. Output
power is decreased at lower mains voltages.
Derating at high ambient temperature
TRAX specification is valid at 23 5C. Max output current times
will be reduced when using TRAX in high ambient temperature.
Derating at lower frequencies
TRAX voltage output specification is at 50 Hz. Maximum voltage
output at lower frequencies is limited by the transformer. Derating
is linear with frequency and max voltage output at 5 Hz is 10% of
rated output.
Measurement accuracy
External 0.05% of reading + 0.05% FS
AC/DC voltage
and current
Internal DC 0.1% of reading + 0.1% FS
current
Internal AC 0.2% of reading + 0.2% FS
current
Internal AC 0.2% of reading + 0.2% FS
voltage
COM
Ethernet port For running the instrument from an external PC
or connect it to an external network.
Connector for For running the instrument wireless from a PC or
Wifi antenna tablet. (Option)
USB 3 USB ports for multipurpose use
Index L
Leakage Reactance....................................... 41
A Leakage reactance measurements................ 27
Action buttons................................................. 45 Load file.......................................................... 46
Analog inputs.................................................. 15 Load Template................................................ 47
Apps............................................................... 20 Logfile............................................................. 48
B M
Binary outputs................................................ 15 Main menu...................................................... 18
C Maintenance................................................... 10
Calculated parameters................................... 24 Manual control................................................ 26
Communication.............................................. 14 Manual Control......................................... 16, 21
Communication ports..................................... 50 Manual test..................................................... 16
Configured test............................................... 16 O
Connecting a device to TRAX........................ 50 Off-line (simulation) mode.............................. 51
Connectors..................................................... 12 Open ground detection..................................... 8
Contact Resistance........................................ 30 Operating instructions.................................... 16
CONTROL...................................................... 15 Oscilloscope................................................... 25
Control knob............................................. 13, 22
P
Controls.......................................................... 12
Power transformer turns-ratio measurements.28
Create a test session in advance................... 16
Product description........................................... 6
CT excitation current...................................... 28
CT ratio with current....................................... 29 R
CT ratio with voltage....................................... 28 Remote control............................................... 50
Reporting........................................................ 44
D
Resistance measurements............................. 26
Data handling................................................. 44
Delete files...................................................... 47 S
Demagnetization............................................ 36 Safety............................................................... 8
DualGround.................................................... 30 Safety instructions............................................ 9
Dual ground measurements........................... 31 Save and report.............................................. 45
Save a test..................................................... 45
E
Short-circuit Impedance................................. 41
EMERGENCY................................................ 14
Short-circuit measurements........................... 27
Excitation current............................................ 26
Side panel...................................................... 12
Excitation Current........................................... 39
Specifications................................................. 54
G STROBE......................................................... 14
Generator settings.......................................... 22 Symbols............................................................ 8
Global settings................................................ 18
T
GUI Settings................................................... 19
Tap changer operation.................................... 32
I Temperature correction.................................. 30
Impedance...................................................... 26 Test object configuration................................. 44
Indicators........................................................ 12 Test object information (nameplate)............... 16
Instrument safety.............................................. 8 TIMING........................................................... 15
Instrument settings......................................... 23 Top panel........................................................ 13
INTERLOCK 1................................................ 14 TRANS........................................................... 15
INTERLOCK 2................................................ 14 Transformer configuration.............................. 33
TRIG IN.......................................................... 15
Turn off TRAX................................................. 18
Turns ratio...................................................... 37
U
Upgrading....................................................... 52
Use measurement as template...................... 17
V
Voltage withstand measurements.................. 29
W
Warning and Caution Notices........................... 8
Warranty........................................................... 7
Warranty repair................................................. 7
Winding resistance......................................... 32
Z
Zero-sequence impedance measurements.... 27
WWW.MEGGER.COM
Subject to change without notice. Printed matter No. ZP-AJ01E Doc. AJ0383DE V04a 2016