CTM 030 PDF
CTM 030 PDF
CTM 030 PDF
2.1 Analytical Range. The analytical range for each gas component
is determined by the electrochemical cell design. A portion of the
analytical range is selected by choosing a span gas concentration
near the flue gas concentrations.
2.1.2 NO2 Span Gas. Choose a span gas concentration such that the
average stack gas reading for each test run is greater than 25
percent of the span gas concentration. Alternatively, choose the
span gas concentration such that it is not greater than the ppm
concentration value of the NO span gas. The tester should be aware
that NO2 cells are generally designed to measure much lower
concentrations than NO cells and the span gas should be chosen
accordingly. If concentration results exceed 125 percent of the span
gas at any time during the sampling run then the test run for that
channel is invalid.
3. DEFINITIONS
3.2 Nominal Range. The range of concentrations over which each cell
is operated (25% to 125% of span gas value). Several nominal ranges
may be used for any given cell as long as the linearity and stability
check results remain within specification.
3.6 Response Time. The amount of time required for the measurement
system to display 95 percent of a step change in gas concentration on
the data recorder.
4.2 Span Calibration Error. Less than 5% of the span gas value for
NO, NO2, and CO channels and less than or equal to 0.5% O2 for the O2
channel.
4.4 Linearity. For the zero, mid-level, and span gases; the
absolute value of the difference between the gas value and the
analyzer response shall not be greater than 2.5% of the span gas
Gas Research Institute Method GRI-96/0008 Revision 7
EMC Conditional Test Method (CTM-030)
October 13, 1997
Page 5
concentration for NO, CO and O2 cells and not greater than 3.0% of
the span gas for NO2 cells.
4.5 Stability Check Response. The analyzer responses to CO, NO, and
NO2 span gases shall not vary more than 2.0% of span gas value over a
30-minute period or more than 1.0 % of the span gas value over a 15-
minute period.
Gas Research Institute Method GRI-96/0008 Revision 7
EMC Conditional Test Method (CTM-030)
October 13, 1997
Page 6
5.1 Measurement System. Use any measurement system that meets the
performance and design specifications in Sections 4 and 5 of this
method. The sampling system shall maintain the gas sample at a
temperature above the dew point up to the moisture removal system.
The sample conditioning system shall be designed so that there are no
entrained water droplets in the gas sample when it contacts the
electrochemical cells. A schematic of an acceptable measurement
system is shown in Figure 1. The essential components of the
measurement system are described below:
5.1.7 Sample Pump. A leak-free pump, to pull the sample gas through
the system at a flow rate sufficient to minimize the response time of
the measurement system. The pump may be constructed of any material
that is nonreactive to the gas being sampled.
5.1.8 Sample Flow Rate Control. A sample flow rate control valve
and rotameter, or equivalent, to maintain a constant sampling rate
within 10 percent during sampling and calibration error checks. The
components shall be fabricated of materials that are nonreactive to
the gas being sampled.
may be used to obtain the analyzer responses and the readings may be
recorded manually.
5.2 Calibration Gases. The calibration gases for the gas analyzer
shall be CO in nitrogen or CO in nitrogen and O2, NO in nitrogen, NO2
in air or nitrogen, and O2 in nitrogen.
5.3.3 Zero Gas. Concentration of less than O.25 percent of the span
gas for each component. Ambient air may be used in a well ventilated
area.
6.2 Linearity Check. Conduct the following procedure once for each
nominal range that is to be used on each electrochemical cell (NO,
NO2, CO, and O2) before each field test program. If a field test
program lasts longer than five days, the linearity check shall be
repeated before each five days of analyzer operation. Repeat the
linearity check if a cell is replaced.
6.2.1 Linearity Check Gases. For each cell obtain the following
gases: zero (0-0.25% of nominal range), mid-level (40-60% of span
gas concentration), and span gas (selected according to Section 2.1).
6.4 Stability Check. Conduct the following procedure once for each
nominal range that is to be used on each pollutant electrochemical
cell (NO, NO2, and CO) before each field test program. If a field
test program lasts longer than five days, the stability check shall
be repeated before each five days of analyzer operation. Repeat the
stability check if a cell is replaced or if a cell is exposed to gas
concentrations greater than 125 percent of the highest span gas
concentration.
6.4.1 Procedure. Inject the span gas into the analyzer and record
the analyzer response at least once per minute until the conclusion
of the test. One-minute average values may be used instead of
instantaneous readings. After the analyzer response has stabilized,
continue to flow the span gas for at least 30 minutes. Make no
adjustments to the analyzer during the test except to maintain
constant flow. Record the stability time as the number of minutes
elapsed between the start of the gas injection and the start of the
30-minute stability check period. If the concentration reaches a
peak value within five minutes, you may choose to record the data for
at least 15 minutes following the peak.
7.2 Warm Up Period. Assemble the sampling system and allow the
analyzer and sample interface to warm up and adjust to ambient
temperature at the location where the stack measurements will take
place.
7.3.2 Inject the zero and span calibration gases at the probe tip
using the calibration assembly. Ensure that the calibration gases
flow through all parts of the sample interface (including any exhaust
lines). During this check, make no adjustments to the system except
those necessary to achieve the correct calibration gas flow rate at
the analyzer. Set the analyzer flow rate to the value recommended by
the analyzer manufacturer. Allow each reading to stabilize before
recording the result on a form similar to Figure 4. The time allowed
for the span gas to stabilize shall be no less than the stability
time noted during the stability check. After achieving a stable
response, disconnect the gas and briefly purge with ambient air.
7.3.5 If the zero and span calibration error test results are not
within the specifications in Section 4, take corrective action and
repeat the calibration error check until acceptable performance is
achieved.
and record the temperature regularly (at least once each 5 minutes)
during the sample collection period. If at any time during sampling
the NO cell temperature is 85_F or greater and has increased or
decreased by more than 5_F since the pretest calibration, stop
sampling immediately and conduct a post test calibration error check
per Section 7.6, re-zero the analyzer, and then conduct another
pretest calibration error check before continuing.
7.6 Post Test Calibration Error Check. Immediately after the test
run or set of test runs conduct span and zero calibration error
checks using the procedure in Section 7.3. Conduct the calibration
error check at the sampling location. Make no changes to the
sampling system or analyzer calibration until all of the calibration
error test results have been recorded. If the zero or span
calibration error exceeds the specifications in Section 4 then all
test data collected since the previous calibration error check are
invalid. If the sampling system is disassembled or the analyzer
calibration is adjusted, repeat the calibration error check before
conducting the next test or test run.
Gas Research Institute Method GRI-96/0008 Revision 7
EMC Conditional Test Method (CTM-030)
October 13, 1997
Page 15
7.8 Re-Zero. At least once every three hours or each time the
analyzer sampling location changes, recalibrate the analyzer at the
zero level according to the manufacturer*s instructions and conduct a
pretest calibration error test before resuming sampling. If the
analyzer is capable of reporting negative concentration data (at
least 5% of the span gas below zero), then the tester is not required
to re-zero the analyzer.
8. CALIBRATION CORRECTIONS
8.1 The tester may choose to correct the emissions data for a test
run using the pretest and post test calibration error results. Use
the following formula to make the corrections:
CMA
CGAS = (CR - Co) _________
CM - Co
where: CGAS = corrected flue gas concentration
CR = flue gas concentration indicated by gas analyzer
Co = average of initial and final zero checks
CM = average of initial and final span checks
CMA = actual concentration of span gas
9. BIBLIOGRAPHY
FIGURE 1.
CALIBRATION SYSTEM SCHEMATIC
Gas Research Institute Method GRI-96/0008 Revision 7
EMC Conditional Test Method (CTM-030)
October 13, 1997
Page 17
FIGURE 2.
LINEARITY CHECK DATA SHEET
Date______________________ Analyst__________________________
FIGURE 3.
STABILITY CHECK DATA SHEET
Date______________________ Analyst__________________________
Stability Time________________
Gas Research Institute Method GRI-96/0008 Revision 7
EMC Conditional Test Method (CTM-030)
October 13, 1997
Page 19
FIGURE 4.
CALIBRATION ERROR CHECK DATA SHEET
Emission Source______________________________________________________
Date__________________
Analyst__________________________