User Manual WINOXLR 3-6-14 Products
User Manual WINOXLR 3-6-14 Products
User Manual WINOXLR 3-6-14 Products
WDESK-R/L
WINOX-R/L/2L
WTAB-R/L
3/6/14 Products
2004/108/EC
SYSTEM IDENTIFICATION
KEY TO SYMBOLS
Below are the symbols used in the manual to draw the reader's attention:
Further information.
GUARANTEE
24 months from the delivery document date. The guarantee covers only defected parts and includes the replacement parts and
labour. All shipping and packing costs are paid by the customer. It is possible to have the repair in guarantee on condition that the
returned product has not been transformed, damaged or repaired without authorization. No guarantee is applicable on returned
products without the original label and/or serial number. No guarantee against misuse.
Batteries: Laumas provides 1 year guarantee from the date of delivery note, against material defects or battery manufacturing faults.
The terminals indicated on the instruments wiring diagram to be connected to earth must
have the same potential as the weighed structure (same earthing pit or earthing system). If
you are unable to ensure this condition, connect with an earthing wire the terminals of the
instrument (including the terminal SUPPLY) to the weighed structure.
The cell cable must be individually led to its panel input and not share a conduit with other cables;
connect it directly to the instrument terminal strip without breaking its route with support terminal
strips.
Use RC filters on the instrument-driven solenoid valve and remote control switch coils.
Avoid inverters in the instrument panel; if inevitable, use special filters for the inverters and separate
them with sheet metal partitions.
The panel installer must provide electric protections for the instruments (fuses, door lock switch
etc.).
It is advisable to leave the equipment always switched on to prevent the formation of condensation.
INSTALLING LOAD CELLS: The load cells must be placed on rigid, stable in-line structures; it
is important to use the mounting modules for load cells to compensate for misalignment of the
support surfaces.
PROTECTION OF THE CELL CABLE: Use water-proof sheaths and joints in order to protect the
cables of the cells.
MECHANICAL RESTRAINTS (pipes, etc.): When pipes are present, we recommend the use of
hoses and flexible couplings with open mouthpieces with rubber protection; in case of hard pipes,
place the pipe support or anchor bracket as far as possible from the weighed structure (at a distance
at least 40 times the diameter of the pipe).
-1-
CONNECTING SEVERAL CELLS IN PARALLEL: Connect several cells in parallel by using - if
necessary - a watertight junction box with terminal box. The cell connection extension cables must
be shielded, led individually into their piping or conduit and laid as far as possible from the power
cables (in case of 4-wire connections, use cables with 4x1 mm2 minimum cross-section).
WELDING: Avoid welding with the load cells already installed. If this cannot be avoided, place the
welder ground clamp close to the required welding point to prevent sending current through the load
cell body.
EARTHING THE WEIGHED STRUCTURE: By means of a copper wire with suitable cross-section,
connect the cell upper support plate with the lower support plate, then connect all the lower plates to
a single earthing system. Electrostatic charges accumulated because of the product rubbing against
the pipes and the weighed container walls are discharged to the ground without going through or
damaging the load cells. Failure to implement a proper earthing system might not affect the
operation of the weighing system; this, however, does not rule out the possibility that the cells and
connected instrument may become damaged in the future. It is forbidden to ensure earthing system
continuity by using metal parts contained in the weighed structure.
OK OK NO
NO NO OK
-2-
LOAD CELL INPUT TEST (QUICK ACCESS)
From the weight display, press for 3 seconds; the response signal of the load cells is displayed,
expressed in mV with four decimals.
- Disconnect the load cells from the instrument and check that there is no moisture in the cell
junction box caused by condensation or water infiltration. If so, drain the system or replace it if
necessary.
- The value between the positive signal wire and the negative signal wire must be equal or similar
to the one indicated in the load cell data sheet (output resistance).
- The value between the positive excitation wire and the negative excitation wire must be equal or
similar to the one indicated in the load cell data sheet (input resistance).
- The insulation value between the shield and any other cell wire and between any other cell wire
and the body of the load cell must be higher than 20 Mohm.
- Take out the load cell to be tested from underneath the container, or alternatively, lift the
container support.
- Make sure that the excitation of two wires of the load cell connected to the instrument (or
amplifier) is 5 Vdc 3%.
- Measure the response signal between the positive and the negative signal wires by directly
connecting them to the tester, and make sure that it is comprised between 0 and 0.5 mV.
- Apply load to the cell and make sure that there is a signal increment.
IF ONE OF THE ABOVE CONDITIONS IS NOT MET, PLEASE CONTACT THE TECHNICAL
ASSISTANCE SERVICE.
-3-
MAIN SPECIFICATIONS OF THE INSTRUMENT
Indicator with 6-wire load cell input installable on table, panel front, wall or column; 6-key membrane
keypad (WTAB: 8-key) with buzzer, real-time clock/calendar with buffer battery.
Two serial ports (RS485 and RS232) for connection to: PC/PLC up to 32 instruments (max 99 with
line repeaters) by ASCII Laumas or ModBus R.T.U. protocol, remote display, printer. Optional:
integrated Profibus DP, DeviceNet, CANopen, Profinet IO, Ethernet/IP, Ethernet TCP/IP,
Modbus/TCP output.
6 PRODUCTS versions are supplied complete with a RELE6PROD 8-relay module (80x160x60 mm,
available with 12/24 VDC or 115 VAC or 230 VAC power supply).
14 PRODUCTS versions are supplied complete with a RELE6PROD 8-relay module
(80x160x60 mm, available with 12/24 VDC or 115 VAC or 230 VAC power supply) and a
RELE14PROD 8-relay module (80x120x60 mm).
Instruments with P, D, N type connectors: included switching power supply plug 24 V 450 mA, input
100240 VAC, 3 meters long cable.
Display:
LED or signalling
Model Display Digit height
symbols
WDESK-R
Red LED, 6 digits, 7 segments,
WINOX-R 20 mm 16
semi-alphanumeric
WTAB-R
WDESK-L
Backlit LCD, 6 digits, 7 segments,
WINOX-L 20 mm max 46
semi-alphanumeric
WTAB-L
Large backlit LCD, 6 digits, 7 segments,
WINOX-2L 40 mm max 46
semi-alphanumeric
Dimensions:
WDESK Version Max. encumbrance Drilling
P - PG9 cable gland
122x226x164 mm
IP67 protection rating 96x186 mm
(connectors included)
Power supply included
-4-
D D-Sub tray
122x226x189 mm
IP40 protection rating 96x186 mm
(connectors included)
Power supply included
D D-Sub tray
(table) 206x286x85 mm
IP40 protection rating (connectors included)
Power supply included
-5-
N IP65 circular connectors
206x286x160 mm
IP65 protection rating 160x248 mm
(connectors included)
Power supply included
D D-Sub tray
BUFFER BATTERY
The instrument is equipped with an internal battery that allows to keep active the internal clock even
in the event of power failure.
At the first start and after long periods of inactivity, leave the instrument on for at least 12
hours to fully charge the battery.
-6-
AFTER A BLACKOUT
After a blackout the instrument DOES NOT come on again automatically, you have to press ON.
To guarantee an automatic restart after a blackout, disable the ON key as follows:
-7-
TECHNICAL SPECIFICATIONS
-8-
ELECTRICAL CONNECTIONS
BASIC INFORMATION
- It is recommended that the power supply negative pole be grounded (WDESK-D, WINOX,
WTAB: connect the earthing system to the dedicated external terminal ).
- It is possible to supply up to eight 350 ohm load cells or sixteen 700 ohm load cells.
- For 4-wire load cells, make a jumper between EX- and REF- and between EX+ and REF+.
- Connect terminal SUPPLY to the RS485 common of the connected instruments in the event
that these receive alternating current input or that they have an optoisolated RS485.
- In case of an RS485 network with several devices it is recommended to activate the 120 ohm
termination resistance on the two devices located at the ends of the network, as described in
section RS485 SERIAL CONNECTION.
- Option E/EC: selects the first 12 formulas.
WARNING! WDESK
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 115 V / 230 V OPTIONS
(R1) PROD. 1
to E-EC OPT.
OUT COMMON
IN COMMON
(R5) SLOW/ALARM
TXD
EXCITATION +
SUPPLY +
RXD
(R2) PROD. 2
REF./SENSE +
SIGNAL +
IN 3
(R3) PROD. 3
STOP
-
SIGNAL -
SUPPLY -
REF./SENSE -
EXCITATION -
L N 1 7 8
SUPPLY OUT +
SUPPLY OUT -
INPUT
supply
5-24 Vdc
to IN COMMMON
RS485 RS232
+ - 115/230 24 VDC
12-24 Vdc OUTPUTS INPUTS LOAD CELLS VAC OUTPUT
supply max 115 Vac supply 6-WIRE load cell
150 mA 5-24 Vdc connection
EXCIT. +
SIGNAL +
EXCIT. -
SIGNAL -
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16
EC OPTION
8 4
6 7
4
5 8
9 to instrument
Current output: 3 10
) The IN3 input can have the following (1) If the analog output is present (ANALOG OUTPUT
functions: OPTION) the following are no longer available:
- APPROVAL (default) - IN3 input
- SEMI-AUTOMATIC ZERO - SLOW/ALARM output
- NET/GROSS WEIGHT - E/EC options
WARNING: connect power supply specified on the plate found on the back of the instrument.
In 115 V and 230 V versions, terminals + SUPPLY and SUPPLY generate continuous voltage
at 24 Vdc only to be used as power supply for instrument inputs.
-9-
E OPTION RELE6PROD24V
12-24 +
Vdc -
L N 3
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24
-
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 IN COM
+
R1
R2
R3
R4
IN1
IN 3
TXD
RXD
STOP
IN2
IN3
4 5 6 7
SIGNAL -
SIGNAL +
SUPPLY -
Buttons not included in the supply
SUPPLY +
IN4
IN COMMON
to E-EC OPT.
ALARM R5
REF./SENSE -
EXCITATION -
EC OPTION
REF./SENSE +
EXCITATION +
OUT COMMON
8 4
6 7 PROD.1 NO
5 8
to instrument
5-24 Vdc
4 9
RELAY
P1
3 10 PROD.1 COM
INPUT supply
2 11
POWER
1 12
8 4 RS485 RS232
PROD.2 NO
+ - PROD.2 COM
RELAY
P2
to IN COMMMON
5-24 Vdc connection PROD.4 NO
P3
PROD.4 COM
9 10 11 12 13 14 15 16
2 3 4
RELAY
P4
PROD.5 COM
PROD.6 NO
- COMMON
+ 0-10 VDC
PROD.6 COM
RELAY
P6
+ 0-20 4-20 mA
CYCLE END NO
- 10 -
CYCLE END COM
EXCIT. -
EXCIT. +
SIGNAL -
SIGNAL +
Current output:
RELAY
SLOW NO
FC
Voltage output:
min load 10 kohm
RELAY
LE
WARNING: connect power supply specified on the plate found on the back of the
SUPPLY OUT +
L N 0 I 0 0 PROD. 6
instrument. In 115 V and 230 V versions, terminals + SUPPLY and SUPPLY
115/230 115/230 24 VDC I 0 0 0 CYCLE END
generate continuous voltage at 24 Vdc only to be used as power supply for VAC VAC OUTPUT
X X X I SLOW
instrument inputs.
RELE6PROD24V RELE14PROD
12-24 +
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 Vdc - PROD.7 NO
RELAY
L N 3
P7
PROD.7 COM
-
+
IN COM
R1
R2
R3
R4
PROD.8 NO
1 2 3
IN 3
TXD
RXD
IN1
PROD.8 COM
STOP
RELAY
P8
IN2
PROD.9 NO
4 5
SIGNAL -
SIGNAL +
SUPPLY -
IN3
SUPPLY +
4 5 6 7
PROD.9 COM
IN COMMON
IN4
to E-EC OPT.
8
RELAY
PROD.10 NO
REF./SENSE -
P9
EXCITATION -
REF./SENSE +
EXCITATION +
OUT COMMON
PROD.10 COM
5-24 Vdc
PROD.1 NO
RELAY
INPUT supply
RELAY
P1
RS485 RS232 PROD.1 COM
P10
SLOW / ALARM R5
POWER
PROD.2 NO
+ -
PROD.2 COM PROD.11 NO
RELAY
P2
RELAY
PROD.3 NO
supply supply 6-WIRE load cell PROD.3 COM PROD.12 NO
to IN COMMMON
5-24 Vdc connection
RELAY
PROD.4 NO PROD.12 COM
P3
RELAY
P12
POWER
9 10 11 12 13 14 15 16
PROD.13 COM
RELAY
P4
RELAY
PROD.14 NO
P13
PROD.14 COM
9 10 11 12 13 14 15 16
RELAY
RELAY
P5
PROD.5 COM
P14
19 20 21 22 23 24 PROD.6 NO
PROD.6 COM RELAY
P6
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 CYCLE END NO
CYCLE END COM RELE6PROD MODULE
- 11 -
RELAY
FC
EXCIT. -
R1 R2 R3 R4 OUTPUTS
EXCIT. +
SIGNAL -
SIGNAL +
17 18 19 20 21 22 23 24
Buttons not included in the supply I I I 0 PROD. 1
RELAY
LE
EC OPTION 0 I I 0 PROD. 2
8 4
6 7
5 8 I 0 I 0 PROD. 3
4 9 to instrument
3 10
2 11 8 4 0 0 I 0 PROD. 4
1 12
) The IN3 input can have the following functions: I I 0 0 PROD. 5
- APPROVAL (default) WARNING! WDESK 115 V / 230 V OPTIONS 0 I 0 0 PROD. 6
WIRING DIAGRAM: 14 PRODUCTS
0 0 I I PROD. 11
SUPPLY OUT +
Access instrument board by removing the six bottom screws and work on the welding side: join the
red points using a stiff wire.
230 230
115 115
- 12 -
-SUPPLY (12/24 VDC)
RS232, RS485: SHIELD, GND
E/EC OPTION: GND
8 21 +LOAD CELL REF/SENSE
115/230 VAC optional version:
-OUTPUT (24 VDC)
RS232, RS485: SHIELD, GND
E/EC OPTION: GND
9 RS232: TXD 22 -LOAD CELL REF/SENSE
10 RS232: RXD 23 -LOAD CELL SIGNAL (-Sig)
OUTPUT No. 1:
- PRODUCT 1 (3 PROD. version)
11 24 +LOAD CELL SIGNAL (+Sig)
- connect to RELE6PROD module
(6/14 PROD. versions)
OUTPUT No. 2:
- PRODUCT 2 (3 PROD. version)
12 L PHASE (115/230 VAC optional ver.)
- connect to RELE6PROD module
(6/14 PROD. versions)
OUTPUT No. 3:
- PRODUCT 3 (3 PROD. version)
13 N NEUTRAL (115/230 VAC optional ver.)
- connect to RELE6PROD module
(6/14 PROD. versions)
GROUND (115/230 VAC optional ver.)
To access the terminal strip on the WDESK instruments with cable glands, you need to
remove the bottom of the instrument unscrewing the six screws.
Internal Internal
Connector Pin Signal
terminal colour
P1 +SUPPLY (12/24 VDC) 7 red
- 13 -
OUTPUT No.1 (max 24 V):
- PRODUCT 1 (3 PROD. version)
1 11 yellow
- connect to RELE6PROD module
(6/14 PROD. versions)
OUTPUT No.2 (max 24 V):
- PRODUCT 2 (3 PROD. version)
2 12 blue
- connect to RELE6PROD module
(6/14 PROD. versions)
OUTPUT No.3 (max 24 V):
- PRODUCT 3 (3 PROD. version)
3 13 white
- connect to RELE6PROD module
(6/14 PROD. versions)
OUTPUT No.4 (max 24 V):
- CYCLE END (3 PROD. version)
4 14 green
- connect to RELE6PROD module
(6/14 PROD. versions)
OUTPUT No.5 (max 24 V):
D3
- SLOW/ALARM (3/14 PROD. version)
Male
5 - ALARM (6 PROD. version) 3 orange
otherwise:
I/O
Analog output +ANALOG OUTPUT (0-10 V)
E/EC option 6 OUTPUT COMMON (max 24 V) 15 purple
7 INPUT No.1: START (+VDC min 5 V max 24 V) 16 grey
8 INPUT No.2: STOP (+VDC min 5 V max 24 V) 17 pink
INPUT No. 3: selectable
(+VDC min 5 V max 24 V)
9 otherwise: 2 brown
Internal Internal
Connector Pin Signal
terminal colour
C2
1 +SUPPLY (12/24 VDC) 7 red
Standard ver.
12/24 VDC:
Power supply
1 2 2 -SUPPLY (12/24 VDC) 8 black
C3
L PHASE L brown
Optional ver.
115/230 VAC: yellow/
Power supply GROUND ,8
green
L N
N NEUTRAL N blue
E
- 15 -
C5 1 RS485/RS232: SHIELD, GND 8 black
- 16 -
KEY TO RELE6PROD MODULE TERMINALS
Terminal Signal
1 PRODUCT 7 NO
2 PRODUCT 7 COM
3 PRODUCT 8 NO
4 PRODUCT 8 COM
5 PRODUCT 9 NO
6 PRODUCT 9 COM
7 PRODUCT 10 NO
8 PRODUCT 10 COM
9 PRODUCT 11 NO
10 PRODUCT 11 COM
11 PRODUCT 12 NO
12 PRODUCT 12 COM
13 PRODUCT 13 NO
14 PRODUCT 13 COM
15 PRODUCT 14 NO
16 PRODUCT 14 COM
- 17 -
INTRODUCTION TO THE OPERATION
The batching programs for 3/6/14 products are used to realize a mixture of different products, which
are loaded on a single weighing structure.
The instrument is able to load automatically a settable amount for each product, driving the relative
batching organ (even two-speed) through the dedicated contact.
The instrument has the following features:
- Maximum 99 settable formulas (see section FORMULAS PROGRAMMING);
- Formulas programming in fixed or variable steps (see section FORMULAS
PROGRAMMING);
- Batching resume after blackout (see section RESUME BATCHING AFTER A POWER CUT);
- Automatic fall calculation (see section FALL);
- Autotare at batching start (see section AUTOTARE);
- Tolerance error control (see section TOLERANCE);
- Precision batching through slow function (see section SLOW);
- Precision batching through tapping function (see section TAPPING FUNCTION);
- Consumption storage (see section CONSUMPTION);
- Print of batching data (see section PRINT AT CYCLE END);
- Alarm contact (see section ALARM RELAY CLOSURE);
BATCHING START:
- via keypad, setting formula and cycle number to be performed;
- via external contact (see section BATCHING START FROM EXTERNAL CONTACT).
For further information on the batching sequence, see section BATCHING.
CONSUMPTION STORAGE:
The instrument, at the end of every batching, stores the consumed amount for each product.
Consumption value can be viewed directly on instrument display in standby mode.
INSTRUMENT CALIBRATION:
The instrument calibration can be performed both with THEORETICAL CALIBRATION, setting the
instrument full scale and the load cells sensitivity (see section THEORETICAL CALIBRATION), and
with REAL CALIBRATION through SAMPLE WEIGHT (see section REAL CALIBRATION (WITH
SAMPLE WEIGHTS)).
ALARM MANAGEMENT:
If an alarm occurs during the batching, the instrument shows the alarm on the display, closes the
related contact (if enable) and waits for operator intervention to abort or continue the batching.
- 18 -
KEY, LED AND SYMBOLS FUNCTIONS
KEYS
WARNING: MODE and PTARE keys are only available in WTAB instruments.
(press immediately
+
followed by )
- 19 -
LED: WDESK-R, WINOX-R, WTAB-R
LED Function
POWER power supply available
NET net weight (semi-automatic tare or preset tare)
0 zero (deviation from zero not more than 0.25 divisions)
stability
kg unit of measure: kg
g unit of measure: g
W1
W2 not used
W3
INPUT 1 LED lit: input 1 closed
INPUT 2 LED lit: input 2 closed
INPUT 3 LED lit: input 3 closed
P1 LED lit: output 1 closed
P2 LED lit: output 2 closed
P4 LED lit: output 3 closed
P8 LED lit: output 4 closed
R5 LED lit: output 5 closed
Into menus LEDs light up in sequence to indicate that it is not displaying a weight.
- 20 -
SYMBOLS: WDESK-L, WINOX-L, WINOX-2L, WTAB-L
6 7 8 9 10 11
5
WDESK-L 4
12
WINOX-L
3 13
WTAB-L 2
1
16 15 14
6 2 1 7 8 5 4 11 9 10
3 12
WINOX-2L
13
16 15 14
Symbol Function
LED POWER power supply available
1 preset tare enabled
2 gross weight
3 net weight (semi-automatic tare or preset tare)
4 stability
5 zero (deviation from zero not more than +/-0.25 divisions)
6 not used
7 batching in progress
8 value displayed is not a weight
9 not used
10 not used
11 not used
12 not used
13 unit of measure reading
14 the number displayed indicates that the corresponding input is closed
15 the number displayed indicates that the corresponding output is closed
16 not used
- 21 -
MENU MAP
Into menus changes are applied right after pressing the ENTER key (no further confirmation is
required).
SYSTEM PARAMETERS
- 22 -
BATCHING CONSTANTS
- 23 -
INSTRUMENT COMMISSIONING
To turn on the instrument press ON. To turn it off press OFF for about 3 seconds: when
appears release the key.
After a blackout the instrument DOES NOT come on again automatically, you have to press ON.
To guarantee an automatic restart after a blackout, disable the ON key (see section AFTER A
BLACKOUT).
Check that the display shows the weight and that when loading the load cells there is an increase in
weight. If there is not check and verify the connections and correct positioning of the load cells.
- If the instrument has already been theoretical CALIBRATED (plant system identification tag
present on the instrument and on the cover: load cells rated data already entered):
Reset to zero (see section TARE WEIGHT ZERO SETTING)
Check the calibration with sample weights and correct the indicated weight if necessary (see
section REAL CALIBRATION (WITH SAMPLE WEIGHTS)).
- If the instrument HAS NOT BEEN CALIBRATED (missing plant system identification tag)
proceed with calibration:
If load cells data are unknown, follow the procedure in section REAL CALIBRATION (WITH
SAMPLE WEIGHTS)
Enter the rated data of load cells following the procedure given in section THEORETICAL
CALIBRATION
Reset to zero (see section TARE WEIGHT ZERO SETTING)
Check the calibration with sample weights and correct the indicated weight if necessary (see
section REAL CALIBRATION (WITH SAMPLE WEIGHTS)).
- If you use the analog output, set the desired analog output type and the full scale value (see
section ANALOG OUTPUT).
- If you use serial communication, set the related parameters (see section SERIAL
COMMUNICATION SETTING).
- Set instruments clock with current date and time (see section DATE AND TIME SETTING)
- 24 -
PROGRAMMING OF SYSTEM PARAMETERS
From the weight display, press simultaneously keys MENU and ESC to access the parameter setting.
THEORETICAL CALIBRATION
- By modifying the theoretical full scale, the sensitivity or divisions, the real calibration is
cancelled and the theoretical calibration only is considered valid.
- If the theoretical full scale and the recalculated full scale in real calibration (see section
REAL CALIBRATION (WITH SAMPLE WEIGHTS)) are equal, this means that the
calibration currently in use is theoretical; if they are different, the calibration in use is the
real calibration based on sample weights.
- By modifying the theoretical full scale, the sensitivity or divisions and all the systems
parameters containing a weight value will be set to default values.
- 25 -
TARE WEIGHT ZERO SETTING
This menu may also be accessed directly from the weight display, holding down the 0
key for 3 seconds.
Perform this procedure after having set the THEORETICAL CALIBRATION data.
Use this function to set to zero the weight of the empty system after commissioning and then later
on to compensate zero variations due to the presence of product residues.
Procedure:
WARNING: Perform this procedure only if its not possible to reset the weighed structure tare, for
example because it contains product that can not be unloaded.
Set in this parameter the estimated zero value (from 0 to max 999999; default: 0).
After having performed the THEORETICAL CALIBRATION and TARE WEIGHT ZERO
SETTING, this function allows correct calibration to be done using sample weights of known
value and, if necessary, any deviations of the indicated value from the correct value to be
corrected.
Load onto the weighing system a sample weight, which must be at least 50% of the maximum
quantity to be weighed.
By confirming the message the flashing value of the weight currently on the system is
displayed. In this phase all of the LEDs are off. Adjust the value on display by using the arrow keys if
necessary. After confirming, the new set weight will appear with all the LEDs flashing.
After an additional confirmation, the message will be restored and by repeatedly pressing
the key ESC the weight will once again be displayed.
- 26 -
Example: for a system of maximum capacity 1000 kg and 1 kg division, two sample weights are
available, one of 500 kg and the other one of 300 kg. Load both weights onto the system and correct
the indicated weight to 800. Now remove the 300 kg weight, the system must show 500; remove the
500 kg weight, too; the system must read zero. If this does not happen, it means that there is a
mechanical problem affecting the system linearity.
WARNING: identify and correct any mechanical problems before repeating the procedure.
- If theoretical full scale and recalculated full scale in real calibration are equal, it means
that the theoretical calibration is currently in use; otherwise, the real calibration based
on sample weights is in use.
- If the correction made changes the previous full scale for more than 20%, all the
parameters with settable weight values are reset to default values.
- 27 -
The filter enables to stabilise a weight as long as its variations are smaller than the corresponding
response time. It is necessary to set this filter according to the type of application and to the full
scale value set.
ZERO PARAMETERS
(from 0 to max full scale; default: 300; considered decimals: 300 30.0 3.00 0.300): this
parameter indicates the maximum weight value resettable by external contact, keypad or serial
protocol.
(from 0 to max 20% of full scale; default: 0): If at switch-on the weight value is lower than
the value set in this parameter and does not exceed the value, the weight is reset. To
disable this function, set 0.
ZERO TRACKING
(from 1 to 5, default: ): When the weight value is stable and, after a second, it
deviates from zero by a figure in divisions smaller or equal to the figure in divisions set in this
parameter, the weight is set to zero. To disable this function, set .
Example: if the parameter is set to 5 and is set to 2, the weight will be automatically
set to zero for variations smaller than or equal to 10 ( x ).
- 28 -
SETTING UNITS OF MEASURE
: kilograms
: grams
: tons
: pounds
: newtons
: litres
: bars
: atmospheres
: pieces
: newton metres
: kilogram metres
: other generic units of measure not included in the list
If the print function is enabled, the symbol corresponding to the selected unit of measure will be
printed after the measured value.
OUTPUTS
3 PRODUCTS version:
- OUTPUTS 1-3: PRODUCTS 1-3.
- OUTPUT 4: CYCLE END.
6 - 14 PRODUCTS version:
- OUTPUTS 1-4: refer to relay modules decoding tables (see wiring diagrams and section TEST)
3 - 14 PRODUCTS version:
- OUTPUT 5 (Default = ): its possible to select one of the following functions:
- (ALARM): the relay is closed when an alarm is present.
- (SLOW): slow function for a precision batching.
6 PRODUCTS VERSION:
- OUTPUT 5: ALARM (the relay is closed when an alarm is present)
- 29 -
INPUTS
- INPUT 1: START
- INPUT 2: STOP
- INPUT 3 (Default = ): Its possible to select one of the following functions:
- (NET/GROSS): by closing this input for no more than one second, its making an
operation of SEMI-AUTOMATIC TARE and the display will show the net weight. To display
the gross weight again, hold the NET/GROSS input closed for 3 seconds.
- (SEMI-AUTOMATIC ZERO): by closing the input for no more than one second, the
weight is set to zero (see section SEMI-AUTOMATIC ZERO (WEIGHT ZERO-SETTING
FOR SMALL VARIATIONS)).
- (APPROVAL): the instrument starts the batching only after verifying that this input is
closed.
To perform a net operation (SEMI-AUTOMATIC TARE), close the NET/GROSS input or press the
TARE f key for less than 3 seconds. The instrument displays the net weight (just set to zero) and the
NET LED lights up. To display the gross weight again, keep the NET/GROSS input closed or press
TARE for 3 seconds.
This operation can be repeated many times by the operator to allow the loading of several products.
Example:
Put the box on the scale, the display shows the box weight; press TARE, the display shows the net
weight to zero; introduce the product in the box, the display shows the product weight. This
operation can be repeated several times.
While the net weight is displayed, keep pressed to display gross weight. When
the key is released the net weight will be displayed again.
The semi-automatic tare operation is not allowed if the gross weight is zero.
- 30 -
PRESET TARE (SUBTRACTIVE TARE DEVICE)
While the net weight is displayed, keep pressed to display the gross weight. When
the key is released the net weight will be displayed again.
ALL THE SEMI-AUTOMATIC TARE (NET) AND PRESET TARE FUNCTIONS WILL BE
LOST WHEN THE INSTRUMENT IS TURNED OFF.
WDESK, WINOX
It is possible to manually set a preset tare value to be subtracted from the display value
provided that the max weight condition is verified.
By default the instrument shows the last programmed preset tare value: to apply it press and
then ENTER.
After setting the tare value, going back to the weight display, the display shows the net weight
(subtracting the preset tare value) and the NET LED lights up to show that a tare has been entered.
To delete a preset tare and return to gross weight display, hold down TARE for about 3 seconds or
keep the NET/GROSS input (if any) closed for the same length of time (3 seconds). The preset tare
value is set to zero. The NET LED is turned off when the gross weight is displayed once again.
WTAB
It is possible to manually set up to 9 different preset tare values to be subtracted from the
display value provided that the max weight condition is verified.
To apply one of the 9 tares ,enter the corresponding menu, display the set value, change it if
necessary and press ENTER. Going back to the weight display, the display shows the net weight
(subtracting the preset tare value) and the NET LED lights up to show that a tare has been entered
To disable the preset tare and return to gross weight display, hold down PTARE or TARE for about
3 seconds or keep the NET/GROSS input (if any) closed for the same length of time (3 seconds).
The NET LED is turned off when the gross weight is displayed once again.
- 31 -
SEMI-AUTOMATIC ZERO (WEIGHT ZERO-SETTING FOR SMALL VARIATIONS)
By closing the SEMI-AUTOMATIC ZERO input, the weight is set to zero; alternatively, by pressing
the 0 key for less than 3 seconds, the message is displayed for 3 seconds, by
pressing ENTER the weight is set to zero.
This function is only allowed if the weight is lower than the value (see section
RESETTABLE WEIGHT SETTING FOR SMALL WEIGHT CHANGES), otherwise the alarm
appears and the weight is not set to zero.
- : it selects the analog output type (420 mA, 020 mA, 010 V, 05 V, 10 V, 5 V;
default: 420 mA).
For the output 10 V and 5 V the soldered jumper SW1 must be closed:
open the instrument;
locate on the analog board, which is mounted perpendicular to the main board, the
soldered jumper SW1 highlighted in the picture below:
- : choice of a weight followed by the analog output: gross () or net (). If the net
function is not active, the analog output varies according to gross weight.
- : set the weight value for which you wish to obtain the minimum analog output value.
Only set a value different from zero if you wish to limit the analog output range; for
instance: for a full scale value of 10000 kg you require an 4 mA signal at 5000 kg and
20 mA at 10000 kg, in this case, instead of zero, set 5000 kg.
- : set the weight value for which you wish to obtain the maximum analog output value; it
must correspond to the value set in the PLC program (default: calibration full scale). E.g.: if I am
using a 420 mA output and in the PLC program I wish to have 20 mA = 8000 kg, I will set the
parameter to 8000.
- 32 -
- : analog output correction to zero: if necessary adjust the analog output, allowing the PLC
to indicate 0. The sign - can be set for the last digit on the left. E.g.: if I use a 420 mA output
and, with the minimum analog setting, the PLC or tester read 4.1 mA, I must set the parameter to
3.9 to obtain 4.0 on the PLC or tester.
- : correction of analog output to full scale: if necessary permit modification of the analog
output by allowing PLC to indicate the value set in the parameter . E.g. if I am using a
420 mA output with the analog set to full scale and the PLC or tester reads 19.9 mA, I must set
the parameter to 20.1 to get 20.0 on the PLC or tester.
Minimum and maximum values which can be set for zero and full scale corrections:
ANALOG OUTPUT TYPE Minimum Maximum
010 V -0.150 10.200
05 V -0.150 5.500
10 V -10.300 10.200
5 V -5.500 5.500
020 mA -0.200 22.000
420 mA -0.200 22.000
NOTE: the analog output may also be used in the opposite manner, i.e. the weight setting that
corresponds to the analog zero () may be greater than the weight set for the analog full scale
(). The analog output will increase towards full scale as the weight decreases; the analog
output will decrease as the weight increases.
For example:
= 10000 = 0 analog output 010 V
- 33 -
SERIAL COMMUNICATION SETTING
- / : communication port.
For more information about protocols and methods of communication, request the
proper manual to technical assistance.
- 34 -
RS232 SERIAL COMMUNICATION
WDESK - WINOX
2
WTAB 3
RS232 TXD PC
RS232 RXD
-SUPPLY 12-24 VDC 5
DB9-F
12-24 VDC
12-24 VDC
RS485 +
RS485 +
RS485 +
RS485 -
RS485 -
RS485 -
-SUPPLY
-SUPPLY
-SUPPLY
max 500 m
24 Vcc
-
+
VIN TX+ RS485 + RS485 +
PC RS232
3 2 5
If the RS485 network exceeds 100 metres in length or baud-rate over 9600 are
used, two terminating resistors are needed at the ends of the network. Two
120 ohm resistors must be connected between the + and terminals of the
line, on the terminal strip of the furthest instruments. Should there be different
instruments or converters, refer to the specific manuals to determine whether it is
necessary to connect the above-mentioned resistors.
Since a two-wire RS485 output may be used directly on the RS-232 input of a PC or remote display,
it is possible to implement instrument connection to an RS-232 port in the following manner:
INSTRUMENT RS232
RS485 RXD
RS485 + GND
- 35 -
ALARM RELAY CLOSURE
The ALARM realy closing can be enabled or disabled for each of the following alarms: approval
contact ();formula not programmed (); maximum weight exceeded (); minimum
weight (); no increase in weight (); no decrease in weight (); tolerance (); fall
(); the PC did not read the batching data ().
: in presence of alarm, the relay is closed (default)
: the relay is not closed even in the presence of alarm
TEST
- Input Test:
: ensure that for each open input is displayed, is displayed when the input is closed.
- Output Test:
: setting ensure that the corresponding output opens. Setting ensure that the
corresponding output closes.
Refer to the following tables to activate the relay of various products (only 6 and 14 products
versions)
- : back-lighting on;
- : back-lighting off;
- : back-lighting goes off after about one minute of no activity; pressing a key or a weight
change turns it on again.
Selecting the item in the main menu, access is obtained to the date and time display menu.
Pressing ENTER several times scrolls through days - months years and hours minutes; pressing
selects the figure to modify; pressing the figure increases; pressing ENTER you can confirm
and go to the next menu item.
- 37 -
OPERATION SETTINGS
: maximum number of products and steps selectable in formula (example: if the version in
use is the 6 products and the parameter is set to "4", you can set the products from 1 to 4 and at
most 4 steps for each formula).
By changing the value, all formulas are deleted.
It is possible to set:
- from 1 to 3 for 3 products version
- from 1 to 6 for 6 products version
- from 1 to 14 for 14 products version
: Select the switch conditions from the opening of the batched product to the following or to
the CYCLE END closing.
- (Default: ): time set in constants ().
- (Default: ): START input closure or ENTER key pressure.
- (Default: ): stable weight.
INFO MENU
- 38 -
PROGRAMMING OF BATCHING CONSTANTS
From weight display press MENU, then press several times until is displayed and
confirm.
MINIMUM WEIGHT
(from 0 to max full scale; default: 10): minimum weight, value at which the scale is considered
empty. Batching start is only allowed if the weight is lower than this value, during the unloading
phase the CYCLE END contact will be opened when the weight reaches this value and after the
safe emptying time is over.
MAXIMUM WEIGHT
(from 0 to max full scale; default: 0): settable and displayable maximum weight. If the
displayed weight exceeds the maximum weight by 9 divisions, the message is displayed; if
in the formulas programming the weight value set is greater than this value, the message is
displayed and the value will not be stored.
By setting 0, the function is disabled.
(from 0.0 to max 999.9 seconds; default 5.0): time that is necessary for a perfect emptying of
the scale. The instrument waits for this time during the unloading phase (CYCLE END closed), after
reaching the minimum weight and before opening the CYCLE END contact to obtain a perfect
emptying of the scale.
WAITING TIME
(from 0.0 to max 999.9 seconds; default 5.0): time between the end batching of one product
and the following one.
- 39 -
NO PRODUCT LOAD TIME
(from 0.0 to max 999.9 seconds; default: 0.0): this parameter allows the product control
during batching. if there is no product load, the instrument waits for a set duration of time before
activating the alarm .
(from 0.0 to max 999.9 seconds; default: 0.0): this parameter allows the product control
during the unloading phase (cycle end). If there is no product extraction, the instrument waits for a
set duration of time before activating the alarm .
WARNING: The control is only actuated after unloading of the product has started (the weight must
decrease by at least 10 divisions).
FALL
By FALL it means the correction of the product amount in fall after the batching STOP. This amount
is in addition to the product already batched causing inaccuracy. The instrument is able to anticipate
the batching STOP, to reduce this uncertainty, with two possible ways:
- AUTOMATIC fall: the instrument automatically calculates the fall;
- MANUAL fall: the instrument applies the fall set by the operator;
WARNING: Setting a value of WAITING TIME () such that the weight is stable at the end of
the batching, otherwise the update of the AUTOMATIC fall is not correct.
(from 0 to 1; default: 0): its possible to select the automatic or the manual fall.
- = 0: MANUAL fall
- = 1: AUTOMATIC fall
(from 0 to maximum weight; default: 0): in this parameter it is possible to set the fall value for
each product (only if = 0) or to display and/or to modify the fall value calculated
automatically by the instrument (only if = 1).
TOLERANCE
(from 0 to maximum weight; default: 0): Adjustable parameter for each product that defines how
much the batched weight value can deviate from the one set in formula. If the batched weight is
higher or lower than the amount to be batched, for a value greater than the tolerance, the instrument
closes the ALARM contact. Press ENTER to continue the batching. Example: if a PRODUCT = 1000
value is set and a TOLERANCE = 100 value is set, the batched weight must not be lower than 900
- 40 -
or higher than 1100 to allow the instrument to continue the batching process. By setting 0, the
function is disabled.
SLOW
(from 0 to maximum weight; default: 0): adjustable parameter for each product. By slow value
we mean the value that, when subtracted from the weight set, minus the fall value, starts the
batching slow phase (closing the relative contact). If the value set is higher than the weight to be
batched, batching will be all in slow phase. Setting 0 disables this function.
Esempio: If PRODUCT = 100, SLOW = 15 and FALL = 10, the slow phase starts when the weight
reaches 75.
TAPPING FUNCTION
In the event that the batching instrument is not equipped with the speed slow it is possible to use
this function to slow down the product batching in the final phase (through opening and closing
cycles of the PRODUCT contact) and to increase the accuracy. Set in parameter the product
amount you want to batch with the tapping function enabled.
SLOW ON
(from 0 to max 9.9; default: 0): time in which the PRODUCT relay remains closed during
the SLOW phase. By setting 0, the function is disabled.
SLOW OFF
(from 0 to max 9.9; default: 0): time in which the PRODUCT relay remains open during the
SLOW phase. By setting 0, the function is disabled.
AUTOTARE
(from 0 to 1; default: 1): autotare enabling (automatic tare at the batching start).
- : autotare disabled
- : autotare enabled
This operation is possible only if the gross weight is lower than the minimum weight (),
otherwise the alarm message is displayed.
(default ): enabling consumption memory (total batched quantity) for each product.
- : consumption enabled;
- : consumption disabled.
- 41 -
PRINT AT CYCLE END
CHECKING PC PRESENCE
(from 0 to max full scale; Default: ): by setting a value different from zero, the
ALARM/SLOW contact does not switch over in case of alarm/slow, but behaves as a SETPOINT
contact; the relay closes when the weight reaches the value set in this parameter.
- 42 -
FORMULAS PROGRAMMING
Select the formula that you wish to program. It is possible to set max 99 formulas.
An attempt to set a quantity higher than the maximum weight () set in the batching constants,
the message is displayed.
DELETING FORMULAS
Confirm with ENTER the prompt, to delete one formula enter the formula number, to delete
all formulas enter 00 (), confirm with ENTER, you will be prompted to confirm (), to
delete confirm again with ENTER otherwise press ESC to cancel the command.
- 43 -
BATCHING
Note: In case of alarm, the batching may be canceled by pressing the ESC button or by closing the
STOP input.
After having selected the formula and set the desired number of batching cycles, the first batching
cycle starts as follows:
1. The instrument will ensure that:
- the formula has been programmed, otherwise it will display the alarm .
- the amount set in the formula does not exceed the set maximum weight (), otherwise
the message is displayed.
- the input IN3 (if it is set as Approval) is closed, otherwise the message will be
displayed; close the input IN3 to start the batching.
- Only if is set: the gross weight is lower than the minimum weight (), otherwise the
alarm is displayed (it is always possible to force the batching start by pressing the
ENTER key).
- Only if = :
- If the consumed amount exceeds 999000, the message appears for a second.
- If the consumed amount exceeds 999999, it is automatically set to zero.
2. Once the above listed conditions are met, the batching of first product programmed in formula is
started; if the time is set and the product is not loaded for at least 20 divisions during
this time interval, the relative alarm is displayed.
3. The number of the product to be batched (e.g. ) is displayed for 1 second and then the net
weight; the product relay closes;
4. Only if is set: once the value set for that product has been reached, minus the fall and
slow values, the SLOW contact closes (if foreseen); if tapping values have been set, the product
contact carries out opening and closing cycles according to and times;
5. Once the value set for that product has been reached, minus any fall value, the product contact
is opened interrupting product flow;
6. After the contact opening, the system shows the weight preceded by the letter and waits:
- Only if = : The waiting time has elapsed ().
- Only if = : The START input has been closed or the ENTER key has been
pressed.
- Only if = : The weight is stable.
7. If the tolerance () is set and the batched quantity is lower than the set quantity minus this
value, is made only one attempt of finishing re-start, in order to improve the batching precision.
In such a case, the PRODUCT contact is closed and returns to step 5, otherwise the alarm
appears. If instead, the batched quantity is greater than the set amount plus this value, the
alarm appears. Press ENTER to cancel the alarm and go on with the batching.
8. Only if = : consumption of the product just batched is stored;
9. If another product to be batched has been set in the formula, the instrument starts batching the
new product and you return to point 3;
10. If there are no more products to be batched in the formula, the system enters the cycle end
phase:
- 44 -
- The CYCLE END contact is closed;
- The display shows the weight preceded by the letter ;
- If printing is enabled: batching data are printed.
11. At the beginning of the product extraction, if the time is set and the product is not
extracted by at least 20 divisions within this interval of time, the alarm message will be
displayed.
12. The system ends the batching (opening of the CYCLE END contact) only after verifying that:
- The weight is lower than the minimum weight ();
- The safe emptying time has elapsed ();
13. Only if = : the instrument waits for data recording on PC, before being available for a
new batching.
If several batching cycles have been set the instrument starts a new cycle.
It is possible to select the required formula via an external selector switch. After selecting the
formula, close the START external contact for at least 0.5 seconds and, verified the conditions
indicated in section BATCHING, the instrument will perform the selected formula batching.
If there is no the formula selection switch (E/EC options), the latest formula entered via the keypad
will be batched, or its possible to set the required formula in the following menu: press the MENU
key for 3 seconds, will be displayed, set the required formula number using the arrow keys
and confirm with ENTER. At batching start, the set formula will be performed. If 00 is set, the latest
formula set via keypad will be recalled.
If at the end of the batching, the START contact is closed, the same batching
sequence is repeated.
This function allows to batch a single product without having to set a special formula.
From weight displaying press START and select the formula; select the product to batch
(eg.: ) and set its amount; confirm with ENTER to perform the batching of the selected
product.
The instrument updates the consumptions, production and stock values (if present and enabled) and
sends the data to the printer (if enabled).
- 45 -
BATCHING START FOR SINGLE PRODUCT WITH STOP BY KEYBOARD
This function allows to batch a single product, leaving the operator the task to stop the batching.
From weight displaying press START and select the formula; select the product to batch
(eg.: ) and set its amount to 0; confirm with ENTER to close the relay of the selected
product and start the batching; press STOP to stop it.
The instrument updates the consumptions, production and stock values (if present and enabled) and
sends the data to the printer (if enabled).
This function allows to unload all the scale contents until it reaches the minimum weight.
From weight displaying press START, select the formula and select the product;
confirm with ENTER to close the CYCLE END relay and start the scale unloading phase; this ends
after reaching the minimum weight and after the safe emptying time has elapsed.
During the batching the display shows the net weight preceded by the letter .
By pressing the product number is displayed.
By pressing the formula number is displayed, by pressing it again the number of running cycle is
displayed.
BATCHING STOP
- Open the START contact and close the STOP contact to stop the batching. If the START contact
is closed, the alarm is displayed.
- To pause the batching at any time, press STOP, the prompt is displayed. Press ENTER
to resume the batching or press again STOP to cancel it completely.
- 46 -
RESUME BATCHING AFTER A POWER CUT
If a blackout occurs during the batching (unloading phase included), when power comes back
appears: press ENTER to resume batching from the point of interruption, press ESC to
cancel the batching and return to the weight displaying.
CONSUMPTION
If in constants consumption is enabled ( = ), the quantities batched for each product are
stored.
CONSUMPTION DELETION
- 47 -
ALARMS
: (only if = 1) it is displayed if, at batching start, the weight on the scale is higher
than the minimum set in constants (). Press ESC to return to the weight displaying,
press ENTER to cancel the alarm and continue with the batching. If the weight comes
back below to the minimum set, the batching starts.
: it is displayed when at the batching start the APPROVAL input is open (if the input 3 is
set as approval; In 3 = ). Close the input to enable batching or cancel it by pressing
ESC.
: it is displayed if at batching end the weight is different from the value set in formula by a
value higher than tolerance. Press ENTER to cancel the alarm and continue with the
batching.
: it is displayed if at batching start the fall value is higher than the product quantity to be
batched. Press ESC to cancel the alarm and the batching.
: it is displayed if, at batching start, the formula recalled for running is not programmed.
Press the key ESC to quit.
: it is displayed if, at batching start, the formula recalled for running exceeds the maximum
weight. Press the key ESC to quit.
: it means that during batching the STOP key has been pressed, temporarily interrupting
the cycle; press the ENTER key to start it again, or STOP to terminate completely the
batching.
: it indicates that a power failure occurred during batching: press ESC to cancel the
batching or press ENTER to resume batching again from the point of interruption.
: during the batching it indicates that the product is not loaded. It is automatically cancelled
if the product increases. Press STOP twice to cancel the batching.
: during the unload (Cycle End contact closed) it indicates that the product is not extracted.
It is automatically cancelled if the product decreases. Press STOP twice to cancel the
unload.
: it is displayed for 3 seconds if you try to stop the batching (by pressing twice STOP or
closing the related input) when the START input is closed. Open the START input to
cancel the alarm.
: batching is cancelled. Press ESC to quit. If the alarm persists, contact technical
assistance.
: it is displayed when there is a weight alarm and it cancels the current batching. Press
ESC to go back to the weight display and check the alarm.
: the load cell is not connected or is incorrectly connected; the load cell signal exceeds
39 mV; the conversion electronics (AD converter) is malfunctioning; the load cell is a
4wire and there are no jumpers between EX- and REF- and between EX+ and REF+.
: the weight display exceeds 110% of the full scale.
: internal instrument converter failure; check load cell connections, if necessary contact
technical assistance.
: the weight exceeds the maximum weight by 9 divisions.
: maximum displayable value exceeded (value higher than 999999 or lower than -999999).
: weight too high: zero setting not possible.
- 48 -
: this message appears in the sample weight setting, in real calibration, after the fifth
sample weight value has been entered.
: the value set for the parameter is beyond the permitted values; press ESC to quit the
setting mode leaving the previous value unchanged. Examples: a number of decimals is
selected for full scale which exceeds the instrument's display potential; value above the
maximum setting value; the weight value set in sample weight verification does not
match the detected mV increase; the analog output correction goes beyond the permitted
limits.
: lock active on menu item, keypad or display.
: Its not possible to display properly the number because is greater than 999999 or less
than -999999.
: buffer battery low, loss of date and time of Real-Time Clock. Confirm with ENTER to
continue; leave the instrument on for at least 12 hours to charge the battery, if the alarm
persists contact technical assistance.
: an incorrect date has been detected: go into the related menu to check and correct it.
: waiting for START closure or ENTER key pressure to continue the batching.
: PC is not connected.
: PC has not read batching data.
With an alarm the relays open and the analog outputs go to the lowest possible value
according to the following table:
RANGE 020 mA 420 mA 05 V 010 V 10 V 5 V
Output value -0.2 mA 3.5 mA -0.5 V -0.5 V 0V 0V
- 49 -
PRINTING EXAMPLES
If the printer has been set (see section SERIAL COMMUNICATION SETTING), from the weight
display press the PRINT key for less than 3 seconds:
- : prints the displayed weight;
- : prints the constants (minimum weight, maximum weight, etc.);
- : prints one or all of the formulas; press ENTER to display , set the formula
number to be printed or 00 to print them all;
- : prints consumption (only if = , see CONSUMPTION FOR EACH PRODUCT).
BATCHING PRINTOUT
INIT.WEIGHT 10 kg
P01 G 301: 300 kg
P02 N 714: 700 kg
P05 N 1025: 1000 kg
TOT G 2040: 2000 kg
TARE 10 kg
P01 N 301: 300 kg
P02 N 714: 700 kg
P05 N 1025: 1000 kg
TOT G 2050: 2000 kg
CONSTANTS PRINTOUT
-----------------------
W--- 14PROD Addr:01
DATE: 01/10/11 08:30:01
CONSTANTS
MIN 10
MASS 0
TIME SIC 5.0
TIME WAIT 5.0
TIME NLOAD 0.0
TIME NUNLOAD 0.0
FALL MANUAL
- 50 -
TIME SLOWON 0.0
TIME SLOWOF 0.0
AUTOTARE YES
TOTAL NO
PRINT NO
PC NO
SLAVE NO
BLACKOUT MANUAL
FORMULA PRINTOUT
-----------------------
W--- 14PROD Addr:01
DATE: 01/10/11 08:30:01
FORMULA: 01
P01 300 kg
P04 400 kg
P07 1000 kg
P11 400 kg
CONSUMPTION PRINTOUT
-----------------------
W--- 14PROD Addr:01
DATE: 01/10/11 08:30:52
CONSUMPTION
P01 2596 kg
P03 1066 kg
P04 487 kg
TOTAL 4149 kg
FROM: 01/09/11 09:16:45
WEIGHT PRINTOUT
-----------------------
W--- 14PROD Addr:01
DATE: 12/09/11 14:48:12
GROSS 1204 kg
NET 831 kg
TARE 373 kg
- 51 -
RESERVED FOR THE INSTALLER
MENU LOCKING
Through this procedure, its possible to block the access to any menu on the instrument.
Select the menu that you wish to lock:
MENU UNLOCKING
Upon instrument power-on hold down the ESC key until the display shows , then proceed as
follows:
CONSTANTS RESTORE (does not erase the calibration): confirm , use arrow keys to select
, set code 6935 and confirm.
- 52 -
PROGRAM SELECTION: confirm and use the arrow keys to select the desired program:
: basic program, setpoint management only.
: to be used when the loaded weighing system correspond to not loaded cells and vice versa
(product increases while weight on load cells actually decreases).
: weight remote display program with setpoint.
: monoproduct loading program.
: monoproduct unloading program.
: 3 products batching.
: 6 products batching.
: 14 product batching.
: no program.
After confirming the choice of the program (except and ), the user must choose its
approval state among the following possible choices:
*) Contact technical assistance to request the proper manual and the correct procedures for
approval, indicating mandatory hardware code and serial number (see section
INSTRUMENT COMMISSIONING).
If you do not have a specific manual for the newly set program, you can request it
to technical assistance.
Press ESC immediately followed by hold them down for about 5 seconds (this operation is also
possible via the MODBUS and ASCII protocols):
- : no lock.
- : keypad lock: if active, when a key is pressed the message is displayed for 3
seconds.
- : keypad and display lock: if active, the kaypad is locked and the display shows the
instrument model (weight is not displayed); by pressing a key the display shows for 3
seconds.
- 53 -
DECLARATION OF CONFORMITY
Dichiarazione di Dichiariamo che il prodotto al quale la presente dichiarazione si riferisce conforme alle norme di
I
conformit seguito citate.
Declaration of We hereby declare that the product to which this declaration refers conforms with the following
GB
conformity standards.
Declaracin de Manifestamos en la presente que el producto al que se refiere esta declaracin est de acuerdo con
E
conformidad las siguientes normas
Wir erklren hiermit, dass das Produkt, auf das sich diese Erklrung bezieht, mit den nachstehenden
D Konformitts-erklrung
Normen bereinstimmt.
Dclaration de Nous dclarons avec cela responsabilit que le produit, auquel se rapporte la prsente dclaration, est
F
conformit conforme aux normes cites ci-aprs.