Seca Ct3000i User Manual

Download as pdf or txt
Download as pdf or txt
You are on page 1of 50
At a glance
Powered by AI
The document provides information about an ECG device including its technical specifications, safety standards, and available configurations.

The device can store 10 seconds of 12-lead ECG, has a sampling frequency of 1000 Hz and digital resolution of 5μV. It also has filters for muscle tremor and line frequency interference.

The device complies with safety standards IEC 601-1, IEC 601-2-25, and has a safety class of I according to IEC 601-1. It also has a patient leakage current of less than 5μA.

P80User Guide - English

Guia de Operación - Español


Guia de Operação - Português

ESAOTE S.P.A
Via Di Caciolle 15
50127 Firenze
Italy
Tel: +39 055 4229 1
Fax: +39 055 4229 208
Internet Web Site http://www.esaote.com

P80User Guide - English


Guia de Operación - Español
Guia de Operação - Português

ESAOTE S.P.A
Via Di Caciolle 15
50127 Firenze
Italy
Tel: +39 055 4229 1
Fax: +39 055 4229 208
Internet Web Site http://www.esaote.com
P80 User Guide - English
Guia de Operación del P80 - Español
Guia de Operação de P80 - Português

Article Number: 9740440002


Revision History

a 3.1998 d 01.2001
b 9.2000 e 06.2001
c 11.2000 f 07.2001

Associated Documents
Guide to the Interpretation and Measurement Program E/ D
Article Number 9740440008

Distributed by:
ESAOTE S.P.A
Via Di Caciolle 15
50127 Firenze
Italy
TEL: +39 055 4229 1
FAX: +39 055 4229 208

Manufactured by:
SCHILLER AG
Altgasse 68
CH-6341 Baar, Switzerland

93/42/EEC Medical Devices


0123 0123 'Notified Body' TÜV P.S.

SAG ( 2.510189f )

P80 User Guide - English


Guia de Operación del P80 - Español
Guia de Operação de P80 - Português

Article Number: 9740440002


Revision History

a 3.1998 d 01.2001
b 9.2000 e 06.2001
c 11.2000 f 07.2001

Associated Documents
Guide to the Interpretation and Measurement Program E/ D
Article Number 9740440008

Distributed by:
ESAOTE S.P.A
Via Di Caciolle 15
50127 Firenze
Italy
TEL: +39 055 4229 1
FAX: +39 055 4229 208

Manufactured by:
SCHILLER AG
Altgasse 68
CH-6341 Baar, Switzerland

93/42/EEC Medical Devices


0123 0123 'Notified Body' TÜV P.S.

SAG ( 2.510189f )
DECLARATION OF CONFORMITY
Electrocardiograph: ESAOTE P-80 / P80 White

We, the undersigned, hereby declare that the medical device (class II a) specified above conforms with the
Essential Requirements listed in Annex I, of EC Directive 93/42/EEC

This declaration is supported by:

TÜV Product Service GmbH, Management Service, D – 80339 Munich

Certificate of approval No:

Q1Z 01 03 41505 002 DIN EN ISO 9001:2000 / DIN EN 46001:1996

G1 01 03 41505 001 Annex II, Section 3 of the Directive 93/42/EEC Medical Devices

Valid date 02/2004.

Baar (Switzerland), 21.03.2001


0123

Markus Bütler
Quality Assurance Manager

DECLARATION OF CONFORMITY
Electrocardiograph: ESAOTE P-80 / P80 White

We, the undersigned, hereby declare that the medical device (class II a) specified above conforms with the
Essential Requirements listed in Annex I, of EC Directive 93/42/EEC

This declaration is supported by:

TÜV Product Service GmbH, Management Service, D – 80339 Munich

Certificate of approval No:

Q1Z 01 03 41505 002 DIN EN ISO 9001:2000 / DIN EN 46001:1996

G1 01 03 41505 001 Annex II, Section 3 of the Directive 93/42/EEC Medical Devices

Valid date 02/2004.

Baar (Switzerland), 21.03.2001


0123

Markus Bütler
Quality Assurance Manager
Where to Obtain Service and Sales Advice
Donde Obtener Servicio y Asesoramiento Comcercial
Onde obter Assistência Técnica e de Vendas

USA: BIOSOUND Inc.


8000 Castleway Drive
PO Box 50858 Indianapolis
Indiana 46250 - 0858
Tel: +1 (317) 849 1793
Fax: +1 (317) 841 8616

Russia: ESAOTE S.p.a


Moscow Representative Office for CIS countries
Khlebniy per., 27
121069 Moscow
Russia
Tel: +7 (095) 232 1833 - 232 0205
Fax: +7 (095) 232 0205

Italy: ESAOTE S.p.a


Via di Caciolle 15
50127 Firenze
Italy
Tel: +39 055 4229 238
Fax: +39 055 414 899

All other Countries: ESAOTE S.p.a


Via di Caciolle 15
50127 Firenze
Italy
Tel: +39 055 4229 1
Fax: +39 055 4229 208
Internet Web Site http://www.esaote.com

Where to Obtain Service and Sales Advice


Donde Obtener Servicio y Asesoramiento Comcercial
Onde obter Assistência Técnica e de Vendas

USA: BIOSOUND Inc.


8000 Castleway Drive
PO Box 50858 Indianapolis
Indiana 46250 - 0858
Tel: +1 (317) 849 1793
Fax: +1 (317) 841 8616

Russia: ESAOTE S.p.a


Moscow Representative Office for CIS countries
Khlebniy per., 27
121069 Moscow
Russia
Tel: +7 (095) 232 1833 - 232 0205
Fax: +7 (095) 232 0205

Italy: ESAOTE S.p.a


Via di Caciolle 15
50127 Firenze
Italy
Tel: +39 055 4229 238
Fax: +39 055 414 899

All other Countries: ESAOTE S.p.a


Via di Caciolle 15
50127 Firenze
Italy
Tel: +39 055 4229 1
Fax: +39 055 4229 208
Internet Web Site http://www.esaote.com
P80

User Guide Pages 1 - 44

Guia de Operación Pág. 45 - 88

Guia de Operação Pág. 89 - 132

P80

User Guide Pages 1 - 44

Guia de Operación Pág. 45 - 88

Guia de Operação Pág. 89 - 132


Contents - User Guide P80

ENGLISH
Safety Notices ................................................................................................................. Page 3
Introduction ..................................................................................................................... Page 5
Operation Modes ............................................................................................................. Page 6
Basic Information ............................................................................................................ Page 8
Keyboard ......................................................................................................................... Page 10
General Settings .............................................................................................................. Page 13
Settings for Automatic Mode .......................................................................................... Page 21
Patient Cable Connections .............................................................................................. Page 25
Recording an ECG in Automatic Mode .......................................................................... Page 28
Recording an ECG in Manual Mode ............................................................................... Page 30
Optional Measurement Program (M) .............................................................................. Page 32
Care & Maintenance ....................................................................................................... Page 36
Trouble Shooting ............................................................................................................. Page 39
Ordering Information ...................................................................................................... Page 40
Technical Data................................................................................................................. Page 42

P80 Page 1
07.2001

Contents - User Guide P80

ENGLISH
Safety Notices ................................................................................................................. Page 3
Introduction ..................................................................................................................... Page 5
Operation Modes ............................................................................................................. Page 6
Basic Information ............................................................................................................ Page 8
Keyboard ......................................................................................................................... Page 10
General Settings .............................................................................................................. Page 13
Settings for Automatic Mode .......................................................................................... Page 21
Patient Cable Connections .............................................................................................. Page 25
Recording an ECG in Automatic Mode .......................................................................... Page 28
Recording an ECG in Manual Mode ............................................................................... Page 30
Optional Measurement Program (M) .............................................................................. Page 32
Care & Maintenance ....................................................................................................... Page 36
Trouble Shooting ............................................................................................................. Page 39
Ordering Information ...................................................................................................... Page 40
Technical Data................................................................................................................. Page 42

P80 Page 1
07.2001
Terms of Warranty
The ESAOTE P80 is warranted against defects in material and
manufacture for the duration of one year (as from date of purchase).
Excluded from this guarantee is damage caused by an accident or
as a result of improper handling. The warranty entitles free PHYSICIAN‘S RESPONSIBILITY
replacement of the defective part. Any liability for subsequent
damage is excluded. The warranty is void if unauthorized or THE P80 ELECTROCARDIOGRAPH IS
unqualified persons attempt to make repairs. PROVIDED FOR THE EXCLUSIVE USE OF
QUALIFIED PHYSICIANS OR PERSONNEL
In case of a defect, contact your dealer or the manufacturer. UNDER THEIR DIRECT SUPERVISION. THE
The manufacturer can only be held responsible for the safety, NUMERICAL AND GRAPHICAL RESULTS FROM
reliability, and performance of the apparatus if: A RECORDING MUST BE EXAMINED WITH
* assembly operations, extensions, readjustments, modifications, RESPECT TO THE PATIENTS OVERALL
or repairs are carried out by persons authorized by him, and CLINICAL CONDITION. THE RECORDING
PREPARATION QUALITY AND THE GENERAL
* the P80 and approved attached equipment are used in RECORDED DATA QUALITY, WHICH COULD
accordance with the manufacturers instructions. EFFECT THE REPORT DATA ACCURACY, MUST
ALSO BE TAKEN INTO ACCOUNT.
THERE ARE NO EXPRESS OR IMPLIED WARRANTIES WHICH
EXTEND BEYOND THE WARRANTIES HEREINABOVE SET IT IS THE PHYSICIANS RESPONSIBILITY TO
FORTH. ESAOTE MAKES NO WARRANTY OF MAKE THE DIAGNOSIS OR TO OBTAIN EXPERT
MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE OPINION ON THE RESULTS, AND TO INSTITUTE
WITH RESPECT TO THE PRODUCT OR PARTS THEREOF. CORRECT TREATMENT IF INDICATED.

FEDERAL LAW IN THE USA RESTRICTS THIS


About this Handbook DEVICE TO SALE BY OR ON THE ORDER OF A
The philosophy of the manufacturer is one of continuous PHYSICIAN
improvement. Our aim is to provide the user with the most up-to-
date information and the latest technological developments.
Your suggestions and comments are welcome on all ESAOTE
documentation. Please contact the ESAOTE Cardiology Corporate
Marketing Department.

Page 2 P80
07.2001

Terms of Warranty
The ESAOTE P80 is warranted against defects in material and
manufacture for the duration of one year (as from date of purchase).
Excluded from this guarantee is damage caused by an accident or
as a result of improper handling. The warranty entitles free PHYSICIAN‘S RESPONSIBILITY
replacement of the defective part. Any liability for subsequent
damage is excluded. The warranty is void if unauthorized or THE P80 ELECTROCARDIOGRAPH IS
unqualified persons attempt to make repairs. PROVIDED FOR THE EXCLUSIVE USE OF
QUALIFIED PHYSICIANS OR PERSONNEL
In case of a defect, contact your dealer or the manufacturer. UNDER THEIR DIRECT SUPERVISION. THE
The manufacturer can only be held responsible for the safety, NUMERICAL AND GRAPHICAL RESULTS FROM
reliability, and performance of the apparatus if: A RECORDING MUST BE EXAMINED WITH
* assembly operations, extensions, readjustments, modifications, RESPECT TO THE PATIENTS OVERALL
or repairs are carried out by persons authorized by him, and CLINICAL CONDITION. THE RECORDING
PREPARATION QUALITY AND THE GENERAL
* the P80 and approved attached equipment are used in RECORDED DATA QUALITY, WHICH COULD
accordance with the manufacturers instructions. EFFECT THE REPORT DATA ACCURACY, MUST
ALSO BE TAKEN INTO ACCOUNT.
THERE ARE NO EXPRESS OR IMPLIED WARRANTIES WHICH
EXTEND BEYOND THE WARRANTIES HEREINABOVE SET IT IS THE PHYSICIANS RESPONSIBILITY TO
FORTH. ESAOTE MAKES NO WARRANTY OF MAKE THE DIAGNOSIS OR TO OBTAIN EXPERT
MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE OPINION ON THE RESULTS, AND TO INSTITUTE
WITH RESPECT TO THE PRODUCT OR PARTS THEREOF. CORRECT TREATMENT IF INDICATED.
FEDERAL LAW IN THE USA RESTRICTS THIS
About this Handbook DEVICE TO SALE BY OR ON THE ORDER OF A
The philosophy of the manufacturer is one of continuous PHYSICIAN
improvement. Our aim is to provide the user with the most up-to-
date information and the latest technological developments.
Your suggestions and comments are welcome on all ESAOTE
documentation. Please contact the ESAOTE Cardiology Corporate
Marketing Department.

Page 2 P80
07.2001
Safety Notices

ENGLISH
TO PREVENT ELECTRIC SHOCK DO NOT DISASSEMBLE THE UNIT. NO SERVICEABLE PARTS INSIDE. REFER SERVICING TO
QUALIFIED PERSONNEL ONLY.
DO NOT USE THIS UNIT IN AREAS WHERE THERE IS ANY DANGER OF EXPLOSION OR THE PRESENCE OF FLAMMABLE GASES
SUCH AS ANAESTHETIC AGENTS.
THIS PRODUCT IS NOT DESIGNED FOR STERILE USE.
SWITCH THE UNIT OFF BEFORE CLEANING AND DISCONNECT FROM THE MAINS.
DO NOT, UNDER ANY CIRCUMSTANCES, IMMERSE THE UNIT OR CABLE ASSEMBLIES IN LIQUID.
THE DEVICE MUST ONLY BE OPERATED USING BATTERY POWER IF THE EARTH CONNECTION IS SUSPECT OR IF THE MAINS
LEAD IS DAMAGED OR SUSPECTED OF BEING DAMAGED.
DO NOT USE HIGH TEMPERATURE STERILISATION PROCESSES (SUCH AS AUTOCLAVING). DO NOT USE E-BEAM OR GAMMA
RADIATION STERILISATION.
DO NOT USE SOLVENT CLEANERS
USE ONLY ACCESSORIES AND OTHER PARTS RECOMMENDED OR SUPPLIED BY ESAOTE. USE OF OTHER THAN RECOMMENDED
OR SUPPLIED PARTS MAY RESULT IN INJURY INACCURATE INFORMATION AND/ OR DAMAGE TO THE UNIT.
THE P80 COMPLIES WITH EMC REGULATIONS FOR MEDICAL PRODUCTS WHICH AFFORDS PROTECTION AGAINST EMISSIONS
AND ELECTRICAL INTERFERENCE. HOWEVER SPECIAL CARE MUST BE EXERCISED WHEN THE UNIT IS USED WITH HIGH
FREQUENCY EQUIPMENT.
IT MUST BE ENSURED THAT NEITHER THE PATIENT NOR THE ELECTRODES (INCLUDING THE NEUTRAL ELECTRODE) COME
INTO CONTACT WITH OTHER PERSONS OR CONDUCTING OBJECTS (EVEN IF THESE ARE EARTHED).
THERE IS NO DANGER WHEN USING THE ECG UNIT FOR A PACEMAKER PATIENT OR WITH SIMULTANEOUS USE OF OTHER
ELECTRICAL STIMULATION EQUIPMENT. HOWEVER, THE STIMULATION UNITS SHOULD ONLY BE USED AT A SUFFICIENT
DISTANCE FROM THE ELECTRODES. IN CASE OF DOUBT, THE PATIENT SHOULD BE DISCONNECTED FROM THE RECORDER.
THIS UNIT IS CF CLASSIFIED ACCORDING TO IEC 601-1. THIS MEANS THAT THE PATIENT CONNECTION IS FULLY ISOLATED
AND DEFIBRILLATION PROTECTED. THE MANUFACTURER CAN ONLY GUARANTEE PROTECTION AGAINST DEFIBRILLATION
VOLTAGE HOWEVER, WHEN THE ORIGINAL ESAOTE PATIENT CABLE IS USED.
DO NOT TOUCH THE CASING DURING DEFIBRILLATION
IF THE PATIENT CABLE SHOULD BECOME DEFECTIVE AFTER DEFIBRILLATION, LEAD OFF WILL BE DISPLAYED AND AN
ACOUSTIC ALARM GIVEN

P80 Page 3
07.2001

Safety Notices

ENGLISH
TO PREVENT ELECTRIC SHOCK DO NOT DISASSEMBLE THE UNIT. NO SERVICEABLE PARTS INSIDE. REFER SERVICING TO
QUALIFIED PERSONNEL ONLY.
DO NOT USE THIS UNIT IN AREAS WHERE THERE IS ANY DANGER OF EXPLOSION OR THE PRESENCE OF FLAMMABLE GASES
SUCH AS ANAESTHETIC AGENTS.
THIS PRODUCT IS NOT DESIGNED FOR STERILE USE.
SWITCH THE UNIT OFF BEFORE CLEANING AND DISCONNECT FROM THE MAINS.
DO NOT, UNDER ANY CIRCUMSTANCES, IMMERSE THE UNIT OR CABLE ASSEMBLIES IN LIQUID.
THE DEVICE MUST ONLY BE OPERATED USING BATTERY POWER IF THE EARTH CONNECTION IS SUSPECT OR IF THE MAINS
LEAD IS DAMAGED OR SUSPECTED OF BEING DAMAGED.
DO NOT USE HIGH TEMPERATURE STERILISATION PROCESSES (SUCH AS AUTOCLAVING). DO NOT USE E-BEAM OR GAMMA
RADIATION STERILISATION.
DO NOT USE SOLVENT CLEANERS
USE ONLY ACCESSORIES AND OTHER PARTS RECOMMENDED OR SUPPLIED BY ESAOTE. USE OF OTHER THAN RECOMMENDED
OR SUPPLIED PARTS MAY RESULT IN INJURY INACCURATE INFORMATION AND/ OR DAMAGE TO THE UNIT.
THE P80 COMPLIES WITH EMC REGULATIONS FOR MEDICAL PRODUCTS WHICH AFFORDS PROTECTION AGAINST EMISSIONS
AND ELECTRICAL INTERFERENCE. HOWEVER SPECIAL CARE MUST BE EXERCISED WHEN THE UNIT IS USED WITH HIGH
FREQUENCY EQUIPMENT.
IT MUST BE ENSURED THAT NEITHER THE PATIENT NOR THE ELECTRODES (INCLUDING THE NEUTRAL ELECTRODE) COME
INTO CONTACT WITH OTHER PERSONS OR CONDUCTING OBJECTS (EVEN IF THESE ARE EARTHED).
THERE IS NO DANGER WHEN USING THE ECG UNIT FOR A PACEMAKER PATIENT OR WITH SIMULTANEOUS USE OF OTHER
ELECTRICAL STIMULATION EQUIPMENT. HOWEVER, THE STIMULATION UNITS SHOULD ONLY BE USED AT A SUFFICIENT
DISTANCE FROM THE ELECTRODES. IN CASE OF DOUBT, THE PATIENT SHOULD BE DISCONNECTED FROM THE RECORDER.
THIS UNIT IS CF CLASSIFIED ACCORDING TO IEC 601-1. THIS MEANS THAT THE PATIENT CONNECTION IS FULLY ISOLATED
AND DEFIBRILLATION PROTECTED. THE MANUFACTURER CAN ONLY GUARANTEE PROTECTION AGAINST DEFIBRILLATION
VOLTAGE HOWEVER, WHEN THE ORIGINAL ESAOTE PATIENT CABLE IS USED.
DO NOT TOUCH THE CASING DURING DEFIBRILLATION
IF THE PATIENT CABLE SHOULD BECOME DEFECTIVE AFTER DEFIBRILLATION, LEAD OFF WILL BE DISPLAYED AND AN
ACOUSTIC ALARM GIVEN

P80 Page 3
07.2001
Page 4 P80
07.2001

Page 4 P80
07.2001
Introduction

ENGLISH
Intended Use
The P80 is an electrocardiograph devoted to acquiring and printing resting ECGs obtained non-invasively.
The original ESAOTE S.p.A patient cable and the standard accessories are part of the unit.
The unit receives and processes all ECG signals simultaneously and continuously records the data in a 10
second memory.
If there is any disturbance (i.e. loose electrode or end of paper), an audible alarm is given and the corre-
sponding indicator lamp flashes. As long as the disturbance remains, no ECG can be stored. As soon as the
defect is eliminated, storage starts.
The P80 is a 3-channel ECG recorder with the following features:
• Low weight and compact dimensions.
• Built-in rechargeable battery for mains-independent use.
• Simple one key operation.
• Automatic or manual recording modes.
• Selectable printing formats with integrated quality thermal printer.
• ECG memory for easy copying.
• Choice of up to 10 languages for printing.
• Optional measurement program for accurate analysis of recording.

P80 Page 5
07.2001

Introduction
Intended Use

ENGLISH
The P80 is an electrocardiograph devoted to acquiring and printing resting ECGs obtained non-invasively.
The original ESAOTE S.p.A patient cable and the standard accessories are part of the unit.
The unit receives and processes all ECG signals simultaneously and continuously records the data in a 10
second memory.
If there is any disturbance (i.e. loose electrode or end of paper), an audible alarm is given and the corre-
sponding indicator lamp flashes. As long as the disturbance remains, no ECG can be stored. As soon as the
defect is eliminated, storage starts.
The P80 is a 3-channel ECG recorder with the following features:
• Low weight and compact dimensions.
• Built-in rechargeable battery for mains-independent use.
• Simple one key operation.
• Automatic or manual recording modes.
• Selectable printing formats with integrated quality thermal printer.
• ECG memory for easy copying.
• Choice of up to 10 languages for printing.
• Optional measurement program for accurate analysis of recording.

P80 Page 5
07.2001
Operation Modes

Auto Mode Recording

Automatic Mode
Connect Patient
Automatic Mode provides a printout giving 10 Pages 25 to 27

seconds of ECG recording of all 12 leads in 2 dif-


ferent formats.
The following can be programmed freely in each Switch Unit ON
of the formats before recording:
• Lead Format Settings
Pages 13 to 24

• Chart Speed
• Measurements
• Average Cycles & Markings
For further information see Paragraph "ECG
Recording in Automatic Mode".
Automatic Recording Automatic Recording
in Format 1 in Format 2
Page 28 to 29 Page 28 to 29

12 lead printout

Page 6 P80
07.2001

Operation Modes

Auto Mode Recording

Automatic Mode
Connect Patient
Automatic Mode provides a printout giving 10 Pages 25 to 27

seconds of ECG recording of all 12 leads in 2 dif-


ferent formats.
The following can be programmed freely in each Switch Unit ON
of the formats before recording:
• Lead Format Settings
Pages 13 to 24

• Chart Speed
• Measurements
• Average Cycles & Markings
For further information see Paragraph "ECG
Recording in Automatic Mode".
Automatic Recording Automatic Recording
in Format 1 in Format 2
Page 28 to 29 Page 28 to 29

12 lead printout

Page 6 P80
07.2001
Operation Modes

ENGLISH
Manual Mode Recording

Manual Mode Connect Patient


Pages 25 to 27
Manual Mode provides a real time printout of 3
leads that are selected and indicated on the key-
board.
Switch Unit ON
The following can be freely selected before or
during recording:
Settings
• Lead Group Pages 13 to 24

• Chart Speed Manual Recording


Pages 30 to 31
• Sensitivity
• Myogram Filter Leads,
Speed,
For further information see Paragraph "ECG Re- Sensitivity,
cording in Manual Mode". Filter
Page 30

Continuous printout of 3 leads

P80 Page 7
07.2001

Operation Modes

ENGLISH
Manual Mode Recording

Manual Mode Connect Patient


Pages 25 to 27
Manual Mode provides a real time printout of 3
leads that are selected and indicated on the key-
board.
Switch Unit ON
The following can be freely selected before or
during recording:
Settings
• Lead Group Pages 13 to 24

• Chart Speed Manual Recording


Pages 30 to 31
• Sensitivity
• Myogram Filter Leads,
Speed,
For further information see Paragraph "ECG Re- Sensitivity,
cording in Manual Mode". Filter
Page 30

Continuous printout of 3 leads

P80 Page 7
07.2001
Basic Information

Location
Do not keep or operate the apparatus in a wet, moist, or dusty environment. Also, avoid exposure to direct
sunlight or heat from other sources. Do not allow the unit to come into contact with acidic vapours or liq-
uids, as such contact may cause irreparable damage. The unit should not be placed near X-ray or diathermy
units, large transformers or motors.
The unit must be placed on a flat surface.
The unit should not be operated in areas where there is any danger of explosion.

Power Supply
The unit can either be operated from the built-in rechargeable battery, or from the mains.
The mains connection is on the rear of the unit.
The mains indicator lamp is always lit when the unit is connected to the mains supply.
A battery indicator lamp confirms battery operation. When the battery capacity is limited, the indicator
flashes. To recharge the battery, connect the apparatus to the mains supply by means of the supplied power
cable. A totally discharged battery needs less than 15 hours to be fully recharged (60% in less than 3
hours).
A fully charged battery lasts approximately 2 hours of normal use. The unit can remain connected to the
mains supply without any danger of damage to either the battery or the unit.

Page 8 P80
07.2001

Basic Information

Location
Do not keep or operate the apparatus in a wet, moist, or dusty environment. Also, avoid exposure to direct
sunlight or heat from other sources. Do not allow the unit to come into contact with acidic vapours or liq-
uids, as such contact may cause irreparable damage. The unit should not be placed near X-ray or diathermy
units, large transformers or motors.
The unit must be placed on a flat surface.
The unit should not be operated in areas where there is any danger of explosion.

Power Supply
The unit can either be operated from the built-in rechargeable battery, or from the mains.
The mains connection is on the rear of the unit.
The mains indicator lamp is always lit when the unit is connected to the mains supply.
A battery indicator lamp confirms battery operation. When the battery capacity is limited, the indicator
flashes. To recharge the battery, connect the apparatus to the mains supply by means of the supplied power
cable. A totally discharged battery needs less than 15 hours to be fully recharged (60% in less than 3
hours).
A fully charged battery lasts approximately 2 hours of normal use. The unit can remain connected to the
mains supply without any danger of damage to either the battery or the unit.

Page 8 P80
07.2001
Basic Information

ENGLISH
Switching On and Off

The P80 is switched on by means of the key and off by means of the key.

The unit is switched off after 5 minutes (30 seconds if battery capacity is limited) if no key is pressed and
the patient cable is not connected.

Potential Equalisation
If the P80 is used in conjunction with other patient connected equipment, we recommend that the potential
equalisation stud ( ) on the rear of the unit is connected to the hospital / building common ground with
the yellow/green ground cable.
When working from an emergency vehicle, the vehicle common ground can be used.

P80 Page 9
07.2001

Basic Information

ENGLISH
Switching On and Off

The P80 is switched on by means of the key and off by means of the key.

The unit is switched off after 5 minutes (30 seconds if battery capacity is limited) if no key is pressed and
the patient cable is not connected.

Potential Equalisation
If the P80 is used in conjunction with other patient connected equipment, we recommend that the potential
equalisation stud ( ) on the rear of the unit is connected to the hospital / building common ground with
the yellow/green ground cable.
When working from an emergency vehicle, the vehicle common ground can be used.

P80 Page 9
07.2001
Keyboard

AUTO START COPY mm/mV mm/s OFF ON


I II III 7 8 9
(aVL I -aVR) 1mV

aVR aVL aVF MAN START AUTO GAIN QRS BEEP


(II aVF III) - mm 4 5 6
! INOP
––––
mV

V1 V2 V3 STOP SHIFT FILTER PAPER


0 1 2 3
FILTER
V4 V5 V6

Page 10 P80
07.2001

Keyboard

AUTO START COPY mm/mV mm/s OFF ON


I II III 7 8 9
(aVL I -aVR) 1mV

aVR aVL aVF MAN START AUTO GAIN QRS BEEP


(II aVF III) - mm 4 5 6
! INOP
––––
mV

V1 V2 V3 STOP SHIFT FILTER PAPER


0 1 2 3
FILTER
V4 V5 V6

Page 10 P80
07.2001
Keyboard

ENGLISH
Switch unit on
Cancel or enable QRS beeper

Switch unit off mm


––––
mV
Automatic ECG sensitivity adjustment from 10

Copy stored ECG to 5 mm/mV in order to avoid overlapping


traces in automatic mode only

Start automatic recording ECG sensitivity selector

(5, 10 or 20 mm/mV)
Start manual recording
Chart speed selector

(5, 25 or 50 mm/s) for


Stop recording manual recording only

1mV
Key for 1 mV indication mark on output during
Switch myogram filter (muscle tremor filter) on
manual recording
or off
Key for initiation of setups and selection of sec-
Lead group selector (forward)
ond format for printout

Lead group selector (backward)

P80 Page 11
07.2001

Keyboard

ENGLISH
Switch unit on
Cancel or enable QRS beeper

Switch unit off mm


––––
mV
Automatic ECG sensitivity adjustment from 10

Copy stored ECG to 5 mm/mV in order to avoid overlapping


traces in automatic mode only

Start automatic recording ECG sensitivity selector

(5, 10 or 20 mm/mV)
Start manual recording
Chart speed selector

(5, 25 or 50 mm/s) for


Stop recording manual recording only

1mV
Key for 1 mV indication mark on output during
Switch myogram filter (muscle tremor filter) on
manual recording
or off
Key for initiation of setups and selection of sec-
Lead group selector (forward)
ond format for printout

Lead group selector (backward)

P80 Page 11
07.2001
Keyboard

Mains indicator (lit when mains connected)

Battery lamp (lit when running on battery power - mains not connected)
(blinking when battery capacity is limited)

PAPER Warning lamp for end of paper or paper jam

INOP Warning lamp for loose electrode connection

FILTER Myogram filter (lit when filter ON)

I II III
(aVL I -aVR) Indicator lamp for selected lead group I, II, III (Standard) (Cabrera: aVL, I, -aVR) in

manual mode only

aVR aVL aVF


(II aVF III) Indicator lamp for selected lead group aVR, aVL, aVF (Standard) (Cabrera: II, aVF,

III) in manual mode only

V1 V2 V3 Indicator lamp for selected lead group V1, V2, V3 in manual mode only

V4 V5 V6 Indicator lamp for selected lead group V4, V5, V6 in manual mode only

Page 12 P80
07.2001

Keyboard

Mains indicator (lit when mains connected)

Battery lamp (lit when running on battery power - mains not connected)
(blinking when battery capacity is limited)

PAPER Warning lamp for end of paper or paper jam

INOP Warning lamp for loose electrode connection

FILTER Myogram filter (lit when filter ON)

I II III
(aVL I -aVR) Indicator lamp for selected lead group I, II, III (Standard) (Cabrera: aVL, I, -aVR) in

manual mode only

aVR aVL aVF


(II aVF III) Indicator lamp for selected lead group aVR, aVL, aVF (Standard) (Cabrera: II, aVF,

III) in manual mode only

V1 V2 V3 Indicator lamp for selected lead group V1, V2, V3 in manual mode only

V4 V5 V6 Indicator lamp for selected lead group V4, V5, V6 in manual mode only

Page 12 P80
07.2001
General Settings

ENGLISH
Each parameter is set by means of a code. This code comprises a combination starting with the SHIFT key
followed by a number of keys and is always confirmed with the STOP key.
As soon as the SHIFT key is pressed, the keyboard is dedicated to the programming function.
The setting is remembered and the keyboard is only released for other functions when the STOP key is
pressed. Once the settings have been confirmed, they are stored in the memory even when the unit is
switched off.
As an example, if you want to set the language on your P80 to English, proceed as follows:

0 2 2

Key 0 Key 2 Key 2

Program sequence

Start programming /
CONFIRM SETTING
Switch to numerical keyboard

On the following pages the programmable parameters and the programming sequences are described in de-
tail.

NOTE THAT THE SETTINGS ARE ALWAYS CONFIRMED BY PRESSING THE KEY.

P80 Page 13
07.2001

General Settings
Each parameter is set by means of a code. This code comprises a combination starting with the SHIFT key

ENGLISH
followed by a number of keys and is always confirmed with the STOP key.
As soon as the SHIFT key is pressed, the keyboard is dedicated to the programming function.
The setting is remembered and the keyboard is only released for other functions when the STOP key is
pressed. Once the settings have been confirmed, they are stored in the memory even when the unit is
switched off.
As an example, if you want to set the language on your P80 to English, proceed as follows:

0 2 2

Key 0 Key 2 Key 2

Program sequence

Start programming /
CONFIRM SETTING
Switch to numerical keyboard

On the following pages the programmable parameters and the programming sequences are described in de-
tail.

NOTE THAT THE SETTINGS ARE ALWAYS CONFIRMED BY PRESSING THE KEY.

P80 Page 13
07.2001
General Settings

Defaults
The default settings are as follows:
Default Settings
Setting P80 P80 with Measurement Option
Language English English
Leads Standard (S) Standard (S)

ECG: 25 mm/s, Short (o)


MECG: 2*6
6*2 (50 mm/s)+1
Format 1 ECG: 25 mm/s, Short (o)
Measurements: Suppressed (-)
Marks: Enabled (+)
ECG: 25 mm/s, Long (ooo)
MECG: none (-)
Format 2 ECG: 25 mm/s, Long (ooo)
Measurements: Suppressed (-)
Marks: Enabled (+)
Rhythm Leads V1, II V1, II
Autom. Centering Enabled (+) Enabled (+)
Paper Type Z-Folded Z-Folded
Printout of Signals Sequential Sequential
Baseline Filter Setting 0.05 Hz 0.05 Hz
Mains Filter Setting 50 Hz 50 Hz
Myogram Filter Setting 35 Hz, OFF 35 Hz, OFF

Page 14 P80
07.2001

General Settings

Defaults
The default settings are as follows:
Default Settings
Setting P80 P80 with Measurement Option
Language English English
Leads Standard (S) Standard (S)
ECG: 25 mm/s, Short (o)
MECG: 2*6
6*2 (50 mm/s)+1
Format 1 ECG: 25 mm/s, Short (o)
Measurements: Suppressed (-)
Marks: Enabled (+)
ECG: 25 mm/s, Long (ooo)
MECG: none (-)
Format 2 ECG: 25 mm/s, Long (ooo)
Measurements: Suppressed (-)
Marks: Enabled (+)

Rhythm Leads V1, II V1, II


Autom. Centering Enabled (+) Enabled (+)
Paper Type Z-Folded Z-Folded
Printout of Signals Sequential Sequential
Baseline Filter Setting 0.05 Hz 0.05 Hz
Mains Filter Setting 50 Hz 50 Hz
Myogram Filter Setting 35 Hz, OFF 35 Hz, OFF

Page 14 P80
07.2001
General Settings

ENGLISH
Check Settings
The defined formats and settings can be checked as follows:

Setup Printout
Entry Key Sequence Action
SHIFT 0 1 any number Printout of programmed setting.

A printout of the defined settings will be produced and gives the following information, depending on the
installed software:
- Unit designation (P80), Software option installed (M = Measurement) and Software version
- Serial number (Service use)
- Leads: Standard (S) or Cabrera (C)
- ECG Format: Long (ooo), Short (o) or Suppressed (-)
- MECG: Average cycles as defined in auto ECG recording setup
(e.g. 4 * 3 (25 mm/s))
- Measurements: Enabled (+) or Suppressed (-)
- Marks: Enabled (+) or Suppressed (-)
- Selected Rhythm leads Leads selected for R1, R2 resp.
- Automatic Centring: Enabled (+) or Suppressed (-)
- Paper Type: Z-Folded or Roll
- Printout of signals: Sequential or Simultaneous
- Baseline Filter: 0.05, 0.15 or 0.30 Hz
- Mains Filter: 50, 60 Hz or OFF (-)
- Myogram Filter: 25 or 35 Hz, ON (+) or OFF (-)

P80 Page 15
07.2001

General Settings

ENGLISH
Check Settings
The defined formats and settings can be checked as follows:

Setup Printout
Entry Key Sequence Action
SHIFT 0 1 any number Printout of programmed setting.

A printout of the defined settings will be produced and gives the following information, depending on the
installed software:
- Unit designation (P80), Software option installed (M = Measurement) and Software version
- Serial number (Service use)
- Leads: Standard (S) or Cabrera (C)
- ECG Format: Long (ooo), Short (o) or Suppressed (-)
- MECG: Average cycles as defined in auto ECG recording setup
(e.g. 4 * 3 (25 mm/s))
- Measurements: Enabled (+) or Suppressed (-)
- Marks: Enabled (+) or Suppressed (-)
- Selected Rhythm leads Leads selected for R1, R2 resp.
- Automatic Centring: Enabled (+) or Suppressed (-)
- Paper Type: Z-Folded or Roll
- Printout of signals: Sequential or Simultaneous
- Baseline Filter: 0.05, 0.15 or 0.30 Hz
- Mains Filter: 50, 60 Hz or OFF (-)
- Myogram Filter: 25 or 35 Hz, ON (+) or OFF (-)

P80 Page 15
07.2001
General Settings
To reset the unit to the basic default settings, proceed as follows:

Reset Setup
Entry Key Sequence Action
SHIFT 0 6 any number Reset to default base settings.

Select Language
The language is selected as follows:
Select Language
Entry Key Sequence Language
1 German
2 English
3 French
4 Swedish
5 American
SHIFT 0 2
6 Italian
7 Spanish
8 Portuguese
9 Dutch
0 Russian

Confirm the selection by pressing STOP.

Page 16 P80
07.2001

General Settings
To reset the unit to the basic default settings, proceed as follows:

Reset Setup
Entry Key Sequence Action
SHIFT 0 6 any number Reset to default base settings.

Select Language
The language is selected as follows:
Select Language
Entry Key Sequence Language
1 German
2 English

3 French
4 Swedish
5 American
SHIFT 0 2
6 Italian
7 Spanish
8 Portuguese
9 Dutch
0 Russian

Confirm the selection by pressing STOP.

Page 16 P80
07.2001
General Settings

ENGLISH
There are three different filters which can be set individually as follows:
• Baseline filter
• Mains filter
• Myogram filter
The setting for each filter is given on the setup printout.

Baseline Filter
The digital Baseline filter suppresses excessive baseline drifts
The setting options are as follows:

Baseline Filter
Entry Key Sequence Filter Setting
0 0.05 Hz (Default)
SHIFT 5 1 0.15 Hz
3 0.30 Hz

The set value is the lower limit of the frequency range and is normally set to 0.05 Hz. The settings 0.15 and
0.30 Hz should only be used when absolutely necessary, as the possibility exists that they could affect the
original ECG signal, especially the ST segments.
Confirm the selection by pressing STOP.

P80 Page 17
07.2001

General Settings
There are three different filters which can be set individually as follows:

ENGLISH
• Baseline filter
• Mains filter
• Myogram filter
The setting for each filter is given on the setup printout.

Baseline Filter
The digital Baseline filter suppresses excessive baseline drifts
The setting options are as follows:

Baseline Filter
Entry Key Sequence Filter Setting
0 0.05 Hz (Default)
SHIFT 5 1 0.15 Hz
3 0.30 Hz

The set value is the lower limit of the frequency range and is normally set to 0.05 Hz. The settings 0.15 and
0.30 Hz should only be used when absolutely necessary, as the possibility exists that they could affect the
original ECG signal, especially the ST segments.
Confirm the selection by pressing STOP.

P80 Page 17
07.2001
General Settings

Mains and Myogram Filter


The Mains filter is an adaptive digital interference filter designed to suppress AC interference without at-
tenuating or distorting the ECG.
The Myogram filter suppresses disturbances caused by strong muscle tremor. The set value will be the
new upper limit of the frequency range as soon as the FILTER key is switched on or programmed as de-
fault when the unit is switched on. When the filter is active the 'FILTER' lamp on the unit is lit.
Select the mains frequency and myogram filter settings as follows:
Myogram Filter
Entry Key Sequence Setting
Mains Filter
2 Myogram Filt. 25 Hz
Entry Key Sequence Filter Setting
3 Myogram Filt. 35 Hz
5 Mains Filter 50 Hz
SHIFT 8 Myo. Filt. ON when switching on unit
SHIFT 8 6 Mains Filter 60 Hz 1
(marked on printout with +)
9 Mains Filter off Myo. Filt. OFF when switching on unit
8
(marked on printout with –)

Confirm the selection by pressing STOP.

The myogram filter is switched on and off manually with the key. As the ECG is stored in Auto-

matic mode unfiltered, it is therefore possible to print the stored ECG either with or without passing the
myogram filter. Filter ON is indicated by the control lamp marked 'FILTER'. When the FILTER key is
pressed again, the filter is switched off and the indicator lamp goes out. The cutoff frequency of the myo-
gram filter is set to either 25 or 35 Hz.

Page 18 P80
07.2001

General Settings

Mains and Myogram Filter


The Mains filter is an adaptive digital interference filter designed to suppress AC interference without at-
tenuating or distorting the ECG.
The Myogram filter suppresses disturbances caused by strong muscle tremor. The set value will be the
new upper limit of the frequency range as soon as the FILTER key is switched on or programmed as de-
fault when the unit is switched on. When the filter is active the 'FILTER' lamp on the unit is lit.
Select the mains frequency and myogram filter settings as follows:
Myogram Filter
Entry Key Sequence Setting
Mains Filter
2 Myogram Filt. 25 Hz
Entry Key Sequence Filter Setting
3 Myogram Filt. 35 Hz
5 Mains Filter 50 Hz
SHIFT 8 Myo. Filt. ON when switching on unit
SHIFT 8 6 Mains Filter 60 Hz 1
(marked on printout with +)
9 Mains Filter off Myo. Filt. OFF when switching on unit
8
(marked on printout with –)

Confirm the selection by pressing STOP.

The myogram filter is switched on and off manually with the key. As the ECG is stored in Auto-

matic mode unfiltered, it is therefore possible to print the stored ECG either with or without passing the
myogram filter. Filter ON is indicated by the control lamp marked 'FILTER'. When the FILTER key is
pressed again, the filter is switched off and the indicator lamp goes out. The cutoff frequency of the myo-
gram filter is set to either 25 or 35 Hz.

Page 18 P80
07.2001
General Settings

ENGLISH
Defining Lead Sequence, Printout & Paper
The required settings can be selected as follows:

Sequences, Print & Paper


Entry Key Sequence Definition
1 Standard Lead Sequence
2 Cabrera Lead Sequence

3 Simultaneous Print
4 Sequential Print
SHIFT 7
5 Auto-Centering ON
6 Auto-Centering OFF
7 Z-Fold Paper
8 Paper Roll

Confirm the selection by pressing STOP.


- The selectable lead groups are:

Standard Cabrera
I aVR V1 V4 aVL II V1 V4
II aVL V2 V5 I aVF V2 V5
III aVF V3 V6 -aVR III V3 V6

P80 Page 19
07.2001

General Settings

ENGLISH
Defining Lead Sequence, Printout & Paper
The required settings can be selected as follows:

Sequences, Print & Paper


Entry Key Sequence Definition
1 Standard Lead Sequence
2 Cabrera Lead Sequence

3 Simultaneous Print
4 Sequential Print
SHIFT 7
5 Auto-Centering ON
6 Auto-Centering OFF
7 Z-Fold Paper
8 Paper Roll

Confirm the selection by pressing STOP.


- The selectable lead groups are:

Standard Cabrera
I aVR V1 V4 aVL II V1 V4
II aVL V2 V5 I aVF V2 V5
III aVF V3 V6 -aVR III V3 V6

P80 Page 19
07.2001
General Settings
- The selectable printout forms are:
- Simultaneous All 12 ECG leads are printed as 4 groups of 3 leads. All ECG leads
are printed in the same time segment (in automatic mode only).
- Sequential All 12 ECG leads are printed as 4 groups of 3 leads. Each group is a
contiguous time segment of approximately 2.5 or 5 seconds (in auto-
matic mode only). When the long form is selected for the printout,
the ECG traces are organised as follows:
0s - 5s (I, II, III)
5s - 10s (avR, avL, avF)
0s - 5s (V1, V2, V3)
5s - 10s (V4, V5, V6)
- Auto-Centering ON All ECG traces are centred dynamically for optimal use of paper
width.
- Auto-Centering OFF ECG traces are set to a fixed baseline position and may possibly
overlap.
- The selectable paper types are:
- Z-folded Z-folded paper, 90 mm wide, with black marks
- Roll Paper roll, 90 mm wide, without black marks

Acoustic QRS Indication

The acoustic QRS indication can be switched on or off at any time by pressing the key.

Page 20 P80
07.2001

General Settings
- The selectable printout forms are:
- Simultaneous All 12 ECG leads are printed as 4 groups of 3 leads. All ECG leads
are printed in the same time segment (in automatic mode only).
- Sequential All 12 ECG leads are printed as 4 groups of 3 leads. Each group is a
contiguous time segment of approximately 2.5 or 5 seconds (in auto-
matic mode only). When the long form is selected for the printout,
the ECG traces are organised as follows:
0s - 5s (I, II, III)
5s - 10s (avR, avL, avF)
0s - 5s (V1, V2, V3)
5s - 10s (V4, V5, V6)
- Auto-Centering ON All ECG traces are centred dynamically for optimal use of paper
width.
- Auto-Centering OFF ECG traces are set to a fixed baseline position and may possibly
overlap.
- The selectable paper types are:
- Z-folded Z-folded paper, 90 mm wide, with black marks
- Roll Paper roll, 90 mm wide, without black marks

Acoustic QRS Indication

The acoustic QRS indication can be switched on or off at any time by pressing the key.

Page 20 P80
07.2001
Settings for Automatic Mode

ENGLISH
Two separate formats for the automatic ECG output can be selected as follows:

Automatic ECG Format


Entry Key
Setup Format
Sequence
1 Setup Format 1
SHIFT
2 Setup Format 2

With this the 2 automatic mode formats are defined as detailed in the following pages.
The ECG format is set as follows:

ECG Format
Entry Key Sequence Output Format
5 No leads are printed
6 Leads are printed in short form (4 sheets)
1
SHIFT or 1 7 Leads are printed in long form (8 sheets)
2
8 Chart speed is 25 mm/s
9 Chart speed is 50 mm/s

Confirm the selection by pressing STOP.

P80 Page 21
07.2001

Settings for Automatic Mode


Two separate formats for the automatic ECG output can be selected as follows:

ENGLISH
Automatic ECG Format
Entry Key
Setup Format
Sequence
1 Setup Format 1
SHIFT
2 Setup Format 2

With this the 2 automatic mode formats are defined as detailed in the following pages.
The ECG format is set as follows:

ECG Format
Entry Key Sequence Output Format
5 No leads are printed
6 Leads are printed in short form (4 sheets)
1
SHIFT or 1 7 Leads are printed in long form (8 sheets)
2
8 Chart speed is 25 mm/s
9 Chart speed is 50 mm/s

Confirm the selection by pressing STOP.

P80 Page 21
07.2001
Settings for Automatic Mode
The Average cycles are defined as follows:

Average Cycles (Measurement Option only)


Entry Key Sequence Output Format
5 No average cycles are printed.
4 * 3 (25 mm/s)
6 The average complexes are printed out in four groups of three
leads at a chart speed of 25 mm/s.
4 * 3 (50 mm/s)
1 7 The average complexes are printed out in four groups of three
SHIFT or 2 leads at a chart speed of 50 mm/s.
2
6 * 2 (50 mm/s) + 1 * Rhy
8 The average complexes are printed out in six groups of two leads
with one rhythm lead at a chart speed of 50 mm/s.
112* 12
* 1 (25 mm/s) + 2 * Rhy
9 The average complexes are printed out for all 12 leads individually
with two rhythm leads at a chart speed of 25 mm/s.

Confirm the selection by pressing STOP.


Note: Lead selection for the 2 rhythm leads are defined on page 24.

Page 22 P80
07.2001

Settings for Automatic Mode


The Average cycles are defined as follows:

Average Cycles (Measurement Option only)


Entry Key Sequence Output Format
5 No average cycles are printed.
4 * 3 (25 mm/s)
6 The average complexes are printed out in four groups of three
leads at a chart speed of 25 mm/s.
4 * 3 (50 mm/s)
1 7 The average complexes are printed out in four groups of three
SHIFT or 2 leads at a chart speed of 50 mm/s.
2
6 * 2 (50 mm/s) + 1 * Rhy
8 The average complexes are printed out in six groups of two leads
with one rhythm lead at a chart speed of 50 mm/s.
112* 12
* 1 (25 mm/s) + 2 * Rhy
9 The average complexes are printed out for all 12 leads individually
with two rhythm leads at a chart speed of 25 mm/s.

Confirm the selection by pressing STOP.


Note: Lead selection for the 2 rhythm leads are defined on page 24.

Page 22 P80
07.2001
Settings for Automatic Mode

ENGLISH
To define the measurements and markings proceed as follows:

Measurements and Markings (Measurement Option only)


Entry Key Sequence Output Format
Detailed table of measurement results is omitted.
5 However, the values of electrical axes, intervals, and heart rate are
not suppressed.
1 6 Detailed table of measurement results is printed.
SHIFT or 3
2 Reference markings (beginning and end of P wave and QRS as
7
well as end of T wave) are omitted.
Reference markings (beginning and end of P wave and QRS as
8
well as end of T wave) are added to the ECG cycles.

Confirm the selection by pressing STOP.


To print or suppress the patient data on the printout proceed as follows:

Patient Data (not with Measurement Option)


Entry Key Sequence Output Format
1 5 Patient data page is omitted.
SHIFT or 4
2 6 Patient data page is printed.

Confirm the selection by pressing STOP.

P80 Page 23
07.2001

Settings for Automatic Mode


To define the measurements and markings proceed as follows:

ENGLISH
Measurements and Markings (Measurement Option only)
Entry Key Sequence Output Format
Detailed table of measurement results is omitted.
5 However, the values of electrical axes, intervals, and heart rate are
not suppressed.
1 6 Detailed table of measurement results is printed.
SHIFT or 3
2 Reference markings (beginning and end of P wave and QRS as
7
well as end of T wave) are omitted.
Reference markings (beginning and end of P wave and QRS as
8
well as end of T wave) are added to the ECG cycles.

Confirm the selection by pressing STOP.


To print or suppress the patient data on the printout proceed as follows:

Patient Data (not with Measurement Option)


Entry Key Sequence Output Format
1 5 Patient data page is omitted.
SHIFT or 4
2 6 Patient data page is printed.

Confirm the selection by pressing STOP.

P80 Page 23
07.2001
Settings for Automatic Mode

Selecting Rhythm Leads (Measurement Option only)


The rhythm leads are printed out in the last two average cycle selections. Two separate rhythm leads can be
selected. The following formats can be set:

Rhythm Leads (Measurement Option only)


Entry Key
Setup Lead
Sequence
3 Setup Rhythm Lead 1
SHIFT
4 Setup Rhythm Lead 2

The 2 rhythm leads are defined as follows:

Extremity Leads Precordial Leads


Entry Key Sequence Lead Entry Key Sequence Lead
1 I 1 V1
2 II 2 V2
3 3 III 3 3 V3
SHIFT or 8 SHIFT or 9
4 4 aVR 4 4 V4
5 aVL 5 V5
6 aVF 6 V6

Confirm the selection by pressing STOP.

Page 24 P80
07.2001

Settings for Automatic Mode

Selecting Rhythm Leads (Measurement Option only)


The rhythm leads are printed out in the last two average cycle selections. Two separate rhythm leads can be
selected. The following formats can be set:

Rhythm Leads (Measurement Option only)


Entry Key
Setup Lead
Sequence
3 Setup Rhythm Lead 1
SHIFT
4 Setup Rhythm Lead 2

The 2 rhythm leads are defined as follows:

Extremity Leads Precordial Leads


Entry Key Sequence Lead Entry Key Sequence Lead
1 I 1 V1
2 II 2 V2
3 3 III 3 3 V3
SHIFT or 8 SHIFT or 9
4 4 aVR 4 4 V4
5 aVL 5 V5
6 aVF 6 V6

Confirm the selection by pressing STOP.

Page 24 P80
07.2001
Patient Cable Connections

ENGLISH
The accessory kit of the electrocardiograph includes a 10-lead patient cable. This cable is plugged into the
patient cable socket on the right-hand side of the unit and secured with the two screws.
The P80 is CF rated. The patient connection is fully isolated and defibrillation protected. Protection
against defibrillation voltage is however only ensured, if the original ESAOTE patient cable is used.
Make sure that during ECG recording neither the patient nor the conducting parts of the patient
connection or the electrodes (including the neutral electrode) come into contact with other persons or
conducting objects (even if these are earthed).
The standard accessories include four stainless steel limb electrodes and 6 precordial suction electrodes.
The limb electrodes are first spread with electrode gel and then attached with the respective clamps to the
arm and foot pick-up sites.
The six suction electrodes are also moistened with gel and placed in their corresponding positions.

P80 Page 25
07.2001

Patient Cable Connections


The accessory kit of the electrocardiograph includes a 10-lead patient cable. This cable is plugged into the

ENGLISH
patient cable socket on the right-hand side of the unit and secured with the two screws.
The P80 is CF rated. The patient connection is fully isolated and defibrillation protected. Protection
against defibrillation voltage is however only ensured, if the original ESAOTE patient cable is used.
Make sure that during ECG recording neither the patient nor the conducting parts of the patient
connection or the electrodes (including the neutral electrode) come into contact with other persons or
conducting objects (even if these are earthed).
The standard accessories include four stainless steel limb electrodes and 6 precordial suction electrodes.
The limb electrodes are first spread with electrode gel and then attached with the respective clamps to the
arm and foot pick-up sites.
The six suction electrodes are also moistened with gel and placed in their corresponding positions.

P80 Page 25
07.2001
Patient Cable Connections
The quality of the ECG is dependent on the preparation and the resistance between the skin and the elec-
trode. To ensure a good quality ECG and minimise the skin/electrode resistance, remember the following
points:
1. Ensure that the patient is warm and relaxed.
2. Shave electrode area before cleaning.
3. Thoroughly clean the area with alcohol.
4. When applying the electrodes ensure that a layer of gel is between the electrode and the skin.
5. Place the C4 electrode first - in the fifth intercostal space on midclavicular line.
Then place:
• C1 in fourth intercostal space at the right sternal border
• C2 in fourth intercostal space at the left sternal border
• C3 between, and equidistant to, C4 and C2
• C6 on left midaxillary line on the same level as C4
• C5 between, and equidistant to, C4 and C6
Following these simple guidelines will ensure good results every time.

Page 26 P80
07.2001

Patient Cable Connections


The quality of the ECG is dependent on the preparation and the resistance between the skin and the elec-
trode. To ensure a good quality ECG and minimise the skin/electrode resistance, remember the following
points:
1. Ensure that the patient is warm and relaxed.
2. Shave electrode area before cleaning.
3. Thoroughly clean the area with alcohol.
4. When applying the electrodes ensure that a layer of gel is between the electrode and the skin.
5. Place the C4 electrode first - in the fifth intercostal space on midclavicular line.
Then place:
• C1 in fourth intercostal space at the right sternal border
• C2 in fourth intercostal space at the left sternal border
• C3 between, and equidistant to, C4 and C2
• C6 on left midaxillary line on the same level as C4
• C5 between, and equidistant to, C4 and C6
Following these simple guidelines will ensure good results every time.

Page 26 P80
07.2001
Patient Cable Connections

ENGLISH
Standard Leads

C3 - GREEN

1 C4 - BROWN
2
3
4
C1 - RED 5 IEC AHA
C6 - VIOLET

N Black RL Green
R Red RA White

C5 - BLACK
C2 - YELLOW
C1 White/Red V1 Brown/Red
RIGHT ARM - RED LEFT ARM - YELLOW
C2 White/Yellow V2 Brown/Yellow
C3 White/Green V3 Brown/Green
C4 White/Brown V4 Brown/Blue
C5 White/Black V5 Brown/Orange
C6 White/Violet V6 Brown/Violet

L Yellow LA Black
F Green LL Red

RIGHT LEG - BLACK LEFT LEG - GREEN

P80 Page 27
07.2001

Patient Cable Connections

ENGLISH
Standard Leads

C3 - GREEN

1 C4 - BROWN
2
3
4
C1 - RED 5 IEC AHA
C6 - VIOLET

N Black RL Green
R Red RA White

C5 - BLACK
C2 - YELLOW
C1 White/Red V1 Brown/Red
RIGHT ARM - RED LEFT ARM - YELLOW
C2 White/Yellow V2 Brown/Yellow
C3 White/Green V3 Brown/Green
C4 White/Brown V4 Brown/Blue
C5 White/Black V5 Brown/Orange
C6 White/Violet V6 Brown/Violet

L Yellow LA Black
F Green LL Red

RIGHT LEG - BLACK LEFT LEG - GREEN

P80 Page 27
07.2001
Recording an ECG in Automatic Mode
In automatic mode, a full 12-lead ECG is printed in one of two predefined formats with a sensitivity of 10
mm/mV. These two formats are selected by the user to suit his specific needs and requirements (as detailed
mm
––––
mV
previously). When the key is pressed before recording in automatic mode, the unit detects very

large waveform amplitudes and sets the sensitivity for the extremity and/or precordial leads to 5 mm/mV to
reduce the overlapping of traces.

=> To start the automatic ECG recording in Format 1, simply press .

=> To start the automatic recording in the second format, press followed by .

The printout provides you with the following information:


- ECG recording of all leads in either Standard or Cabrera format according to selection
- Sensitivity
- Heart Rate
- Speed
and if set:
- Patient Data field to manually insert patient data (Standard version only)
- Average Cycles (M version only)
- Intervals (M version only)
- Axis (M version only)
- Detailed Measurement Table (M version only)

Page 28 P80
07.2001

Recording an ECG in Automatic Mode


In automatic mode, a full 12-lead ECG is printed in one of two predefined formats with a sensitivity of 10
mm/mV. These two formats are selected by the user to suit his specific needs and requirements (as detailed
mm
––––
mV
previously). When the key is pressed before recording in automatic mode, the unit detects very

large waveform amplitudes and sets the sensitivity for the extremity and/or precordial leads to 5 mm/mV to
reduce the overlapping of traces.

=> To start the automatic ECG recording in Format 1, simply press .

=> To start the automatic recording in the second format, press followed by .

The printout provides you with the following information:


- ECG recording of all leads in either Standard or Cabrera format according to selection
- Sensitivity
- Heart Rate
- Speed
and if set:
- Patient Data field to manually insert patient data (Standard version only)
- Average Cycles (M version only)
- Intervals (M version only)
- Axis (M version only)
- Detailed Measurement Table (M version only)

Page 28 P80
07.2001
Recording an ECG in Automatic Mode

ENGLISH
• To obtain an extra printout of the ECG recording in Format 1, simply press .

• To obtain an extra printout of the second format, press followed by .

P80 Page 29
07.2001

Recording an ECG in Automatic Mode

ENGLISH
• To obtain an extra printout of the ECG recording in Format 1, simply press .

• To obtain an extra printout of the second format, press followed by .

P80 Page 29
07.2001
Recording an ECG in Manual Mode
Manual mode provides a direct printout of the real-time ECG with full control of parameter selection.
The following can be freely chosen during or before the recording:

• Lead Group (by means of the and keys)

The following lead groups are selectable:


- I, II, III (Cabrera: aVL, I, -aVR)
- aVR, aVL, aVF (Cabrera: II, aVF, III)
- V1, V2, V3
- V4, V5, V6

- V1, II, V5 by pressing and

• Chart Speed (by means of the , and keys)

• Sensitivity (by means of the , and keys)

Page 30 P80
07.2001

Recording an ECG in Manual Mode


Manual mode provides a direct printout of the real-time ECG with full control of parameter selection.
The following can be freely chosen during or before the recording:

• Lead Group (by means of the and keys)

The following lead groups are selectable:


- I, II, III (Cabrera: aVL, I, -aVR)
- aVR, aVL, aVF (Cabrera: II, aVF, III)
- V1, V2, V3
- V4, V5, V6

- V1, II, V5 by pressing and

• Chart Speed (by means of the , and keys)

• Sensitivity (by means of the , and keys)

Page 30 P80
07.2001
Recording an ECG in Manual Mode

ENGLISH
=> To start the manual recording of a real-time ECG, press the key.

The printout provides you with the following information:


- The group of the three selected leads with lead identification.
- On the lower edge chart speed, sensitivity and filter settings (if on) are given.
- At the top, the heart rate as current average of 4 beats is shown.

1mV
To re-centre the ECG traces, press the key during operation.

=> Finish the recording by pressing the key.

WARNING:
After heavy artefacts or lead off, the indication of the heart rate may not be reliable.

P80 Page 31
07.2001

Recording an ECG in Manual Mode

ENGLISH
=> To start the manual recording of a real-time ECG, press the key.

The printout provides you with the following information:


- The group of the three selected leads with lead identification.
- On the lower edge chart speed, sensitivity and filter settings (if on) are given.
- At the top, the heart rate as current average of 4 beats is shown.

1mV
To re-centre the ECG traces, press the key during operation.

=> Finish the recording by pressing the key.

WARNING:
After heavy artefacts or lead off, the indication of the heart rate may not be reliable.

P80 Page 31
07.2001
Optional Measurement Program (M)

Heart rate (HR)


Average heart rate calculated on the basis of the entire recording (10 seconds) and is shown as number of
beats per minute.

Intervals
RR: Average time interval between two consecutive ventricular complexes, computed on the basis of the
average heart rate.
P: Duration of P wave.
PR: P-R interval, i.e. period of time between beginning of P wave and beginning of QRS complex.
QRS: Duration of QRS complex.
QT: Interval between beginning of QRS (beginning of ventricular depolarisation) and end of T wave
(end of repolarisation phase).
QTC: Normalized QT interval. As the QT interval is dependent on the heart rate, it is often converted into
the normalized QTC interval (i.e. the QT of the patient would show at a HR of 60/min). Usually, the
QTC amounts to 390 ±40 msec. The conversion is achieved according to the following formula:
1000
QTC = QT *
RR

Page 32 P80
07.2001

Optional Measurement Program (M)

Heart rate (HR)


Average heart rate calculated on the basis of the entire recording (10 seconds) and is shown as number of
beats per minute.

Intervals
RR: Average time interval between two consecutive ventricular complexes, computed on the basis of the
average heart rate.
P: Duration of P wave.
PR: P-R interval, i.e. period of time between beginning of P wave and beginning of QRS complex.
QRS: Duration of QRS complex.
QT: Interval between beginning of QRS (beginning of ventricular depolarisation) and end of T wave
(end of repolarisation phase).
QTC: Normalized QT interval. As the QT interval is dependent on the heart rate, it is often converted into
the normalized QTC interval (i.e. the QT of the patient would show at a HR of 60/min). Usually, the
QTC amounts to 390 ±40 msec. The conversion is achieved according to the following formula:

QTC = QT * 1000
RR

Page 32 P80
07.2001
Optional Measurement Program (M)

ENGLISH
Electrical axes
The electrical axes of the heart are determined separately for the P, T and QRS waves. They indicate the
main spreading direction of the electrical vector in the frontal plane.
The ESAOTE measurement program calculates the axes on the basis of the maximal deflection of the rel-
evant waves in leads I and aVF. The following formula is used for the calculation:

axis a = arc tan


( (R+R') + (S+S') [aVF]

(R+R') + (S+S') [I] )


Please note that large discrepancies may be found between two measurements if the P and T waves are in-
distinct. It is also a known fact that breathing and the position (supine or standing) of the patient produce
changes in the electrical axes.

Detailed Measurements for Each Lead


The ESAOTE measurement program prints a table with lead-specific measurements results.
In 12 columns, i.e. for each standard lead, the amplitude values of the P, Q, R, S, R’, S’, T, and T’ waves,
the J point and the ST integral are listed in millivolts (mV). The amplitude measurements relate to a refer-
ence value that corresponds to the signal value immediately before the beginning of the QRS complex .
The duration of the Q, R, S, R’ and S’ waves is indicated in milliseconds (ms) (see illustration on next
page).

P80 Page 33
07.2001

Optional Measurement Program (M)

ENGLISH
Electrical axes
The electrical axes of the heart are determined separately for the P, T and QRS waves. They indicate the
main spreading direction of the electrical vector in the frontal plane.
The ESAOTE measurement program calculates the axes on the basis of the maximal deflection of the rel-
evant waves in leads I and aVF. The following formula is used for the calculation:

axis a = arc tan


( (R+R') + (S+S') [aVF]

(R+R') + (S+S') [I] )


Please note that large discrepancies may be found between two measurements if the P and T waves are in-
distinct. It is also a known fact that breathing and the position (supine or standing) of the patient produce
changes in the electrical axes.

Detailed Measurements for Each Lead


The ESAOTE measurement program prints a table with lead-specific measurements results.
In 12 columns, i.e. for each standard lead, the amplitude values of the P, Q, R, S, R’, S’, T, and T’ waves,
the J point and the ST integral are listed in millivolts (mV). The amplitude measurements relate to a refer-
ence value that corresponds to the signal value immediately before the beginning of the QRS complex .
The duration of the Q, R, S, R’ and S’ waves is indicated in milliseconds (ms) (see illustration on next
page).

P80 Page 33
07.2001
Optional Measurement Program (M)

Qd J-point

ST
R
Q S

P T T'

Rd Sd

R'
R
S S'

Rd Sd
R'd S'd

Page 34 P80
07.2001

Optional Measurement Program (M)

Qd J-point

ST
R
Q S

P T'
T

Rd Sd

R'
R
S S'

Rd Sd
R'd S'd

Page 34 P80
07.2001
Optional Measurement Program (M)

ENGLISH
PARAMETER DESCRIPTION MEAS´MT UNIT
P: amplitude of P wave mV
Q: amplitude of Q wave mV
Qd: duration of Q wave ms
R: amplitude of R wave mV
Rd: duration of R wave ms
S: amplitude of S wave mV
Sd: duration of S wave ms
R’: amplitude of R’ wave mV
R’d: duration of R’ wave ms
S’: amplitude of S’ wave mV
S’d: duration of S’ wave ms
J: amplitude of J point mV
ST integral: averaged amplitude of ST segment
ST: (from J point to half the distance between J-point mV
and T wave maximum)
T: amplitude of T wave mV
amplitude of T’ wave (in case of a diphasic T
T’: mV
wave)

P80 Page 35
07.2001

Optional Measurement Program (M)

ENGLISH
PARAMETER DESCRIPTION MEAS´MT UNIT
P: amplitude of P wave mV
Q: amplitude of Q wave mV
Qd: duration of Q wave ms
R: amplitude of R wave mV
Rd: duration of R wave ms
S: amplitude of S wave mV
Sd: duration of S wave ms
R’: amplitude of R’ wave mV
R’d: duration of R’ wave ms
S’: amplitude of S’ wave mV
S’d: duration of S’ wave ms
J: amplitude of J point mV
ST integral: averaged amplitude of ST segment
ST: (from J point to half the distance between J-point mV
and T wave maximum)
T: amplitude of T wave mV
amplitude of T’ wave (in case of a diphasic T
T’: mV
wave)

P80 Page 35
07.2001
Care & Maintenance

Care of your P80


The patient cable should not be exposed to excessive mechanical stress. Whenever disconnecting the leads,
hold the plugs and not the cables. Store the leads in such a way as to prevent anyone stumbling over them
or any damage caused by the wheels of instrument trolleys. The cable can be wiped with soapy water. Steri-
lization, if required, should be done with gas only and not with steam. To disinfect, wipe the cable with any
standard hospital disinfectant.
The electrodes can be cleaned after every use with soapy water. Make sure that no water is left in the suc-
tion cup of the suction electrodes. Sterilization can also be performed with gas or with Alhydex or Vygon.
The casing of the P80 should be cleaned with a soft cloth on the surface only. Disconnect the unit before
cleaning.
Do not, under any circumstances, immerse the apparatus into a cleaning liquid or sterilize with hot
water, steam, or air.

Self-test
Initiate a self-test of the P80 as follows:

Initiate Self-Test
Entry Key Sequence Action
SHIFT 0 3 any number Printout of Self-test

A table giving information for the service staff is printed out.

Page 36 P80
07.2001

Care & Maintenance

Care of your P80


The patient cable should not be exposed to excessive mechanical stress. Whenever disconnecting the leads,
hold the plugs and not the cables. Store the leads in such a way as to prevent anyone stumbling over them
or any damage caused by the wheels of instrument trolleys. The cable can be wiped with soapy water. Steri-
lization, if required, should be done with gas only and not with steam. To disinfect, wipe the cable with any
standard hospital disinfectant.
The electrodes can be cleaned after every use with soapy water. Make sure that no water is left in the suc-
tion cup of the suction electrodes. Sterilization can also be performed with gas or with Alhydex or Vygon.
The casing of the P80 should be cleaned with a soft cloth on the surface only. Disconnect the unit before
cleaning.
Do not, under any circumstances, immerse the apparatus into a cleaning liquid or sterilize with hot
water, steam, or air.

Self-test
Initiate a self-test of the P80 as follows:

Initiate Self-Test
Entry Key Sequence Action
SHIFT 0 3 any number Printout of Self-test

A table giving information for the service staff is printed out.

Page 36 P80
07.2001
Care & Maintenance

ENGLISH
Maintenance
At 12 monthly intervals, the unit should undergo a technical safety check. The extent of this check should
include the following:
- Visual inspection of the unit and cables.
- Electrical safety tests according to IEC 601-1 and IEC 601-2-25.
- Functional tests according to the Service Handbook.
The test results must be documented.

Replacing the Recording Paper


The recording paper must be replaced as soon as the end of the paper is indicated by a red stripe on the
lower edge. After the indication first appears, there are about 8 pages left. However, we recommend that the
paper be replaced immediately.
If no paper is left, the printing process is interrupted and the paper warning lamp starts to blink. After the
paper has been replaced, the printout is started again by pressing COPY or MAN START or AUTO
START.
A step-by-step description of how to change the paper is given on the next page.
NOTE: If changing from Z-folded paper to a paper roll (or vice versa), be sure that the settings are cor-
rected as follows:
Paper Settings
Entry Key Sequence Definition
7 Z-Fold Paper
SHIFT 7
8 Paper Roll

P80 Page 37
07.2001

Care & Maintenance

ENGLISH
Maintenance
At 12 monthly intervals, the unit should undergo a technical safety check. The extent of this check should
include the following:
- Visual inspection of the unit and cables.
- Electrical safety tests according to IEC 601-1 and IEC 601-2-25.
- Functional tests according to the Service Handbook.
The test results must be documented.

Replacing the Recording Paper


The recording paper must be replaced as soon as the end of the paper is indicated by a red stripe on the
lower edge. After the indication first appears, there are about 8 pages left. However, we recommend that the
paper be replaced immediately.
If no paper is left, the printing process is interrupted and the paper warning lamp starts to blink. After the
paper has been replaced, the printout is started again by pressing COPY or MAN START or AUTO
START.
A step-by-step description of how to change the paper is given on the next page.
NOTE: If changing from Z-folded paper to a paper roll (or vice versa), be sure that the settings are cor-
rected as follows:
Paper Settings
Entry Key Sequence Definition
7 Z-Fold Paper
SHIFT 7
8 Paper Roll

P80 Page 37
07.2001
Care & Maintenance
• Push paper compartment release and lift up cover.

PRINTED SIDE PAPER MARK


UPPERMOST

PAPER RELEASE CATCH

SCHILLER

• Remove the remaining paper.


• Place a new paper pack into the drawer.
• Check that the printed (grid) side is facing upwards, and place the beginning of the paper over the guide
roller.
• Close lid and press firmly until release catches.
• Press STOP to transport the paper to the start position.
NOTE: ESAOTE can only guarantee an excellent printout if ESAOTE original chart paper or a chart
paper of the same quality is used.

Page 38 P80
07.2001

Care & Maintenance


• Push paper compartment release and lift up cover.

PRINTED SIDE PAPER MARK


UPPERMOST

PAPER RELEASE CATCH

SCHILLER

• Remove the remaining paper.


• Place a new paper pack into the drawer.
• Check that the printed (grid) side is facing upwards, and place the beginning of the paper over the guide
roller.
• Close lid and press firmly until release catches.
• Press STOP to transport the paper to the start position.
NOTE: ESAOTE can only guarantee an excellent printout if ESAOTE original chart paper or a chart
paper of the same quality is used.
Page 38 P80
07.2001
Trouble Shooting

ENGLISH
Problem What to Check
Unit does not switch On/ Mains Check if mains cable is plugged in.
Indicator Lamp is not lit Call your local ESAOTE dealer if problem is still present.
CHECK ELECTRODE CONTACT. As much as possible, ensure
that patient is relaxed and warm. Activate myogram filter to reduce
"Noisy" traces
muscle tremor. Set mains filter to 50 or 60 Hz according to local
power supply.
ECG trace "wanders" away from
Baseline drift - check electrode contact. Activate baseline filter.
centre
Check if paper compartment is closed tight. Check paper level.
No printout
Check paper settings.

P80 Page 39
07.2001

Trouble Shooting

ENGLISH
Problem What to Check
Unit does not switch On/ Mains Check if mains cable is plugged in.
Indicator Lamp is not lit Call your local ESAOTE dealer if problem is still present.
CHECK ELECTRODE CONTACT. As much as possible, ensure
that patient is relaxed and warm. Activate myogram filter to reduce
"Noisy" traces
muscle tremor. Set mains filter to 50 or 60 Hz according to local
power supply.
ECG trace "wanders" away from
Baseline drift - check electrode contact. Activate baseline filter.
centre
Check if paper compartment is closed tight. Check paper level.
No printout
Check paper settings.

P80 Page 39
07.2001
Ordering Information
Your local representative stocks all the disposables and accessories available for the P80. In case of diffi-
culty or to obtain the address of your local dealer, please contact the head office. Our staff will be pleased
to help process your order or to provide any details for all products.
The address for advice is:
ESAOTE SPA
Service Department
Via di Caciolle 15
I-50127 Florence
Italy
Phone Number: + (39) 55 42 291
Fax Number: + (39) 55 42 29 208

When ordering, state that the order is for a P80 unit and provide the following:
* Part Title
* Part Number
* Contact name and Address

Page 40 P80
07.2001

Ordering Information
Your local representative stocks all the disposables and accessories available for the P80. In case of diffi-
culty or to obtain the address of your local dealer, please contact the head office. Our staff will be pleased
to help process your order or to provide any details for all products.
The address for advice is:
ESAOTE SPA
Service Department
Via di Caciolle 15
I-50127 Florence
Italy
Phone Number: + (39)0 55 42 291
Fax Number: + (39) 055 42 29 208

When ordering, state that the order is for a P80 unit and provide the following:
* Part Title
* Part Number
* Contact name and Address

Page 40 P80
07.2001
Ordering Information

ENGLISH
Complete Units
P80 Basic, 230 V; Complete with standard accessories ....................................................................... 9840400010
P80 Measurements, 230 V; Complete with standard accessories ......................................................... 9840400020
Standard Accessories:
Patient cable EURO colours; pack of paper; set of ten electrodes
ECG gel; User Guide
P80 Basic, 115 V; Complete with standard accessories ....................................................................... 9840400011
P80 Measurement, 115 V; Complete with standard accessories .......................................................... 9840400021
Standard Accessories:
Patient cable USA colours; pack of paper; set of ten electrodes
ECG gel; User Guide

Accessories
- 10-lead Patient Cable, Euro ............................................................................................................. 9740410001
- 10-lead Patient Cable, USA ............................................................................................................. 9740410002
- 4 Extremity clamp-electrodes .......................................................................................................... 9610020000
- 4 Extremity clamp-electrodes, USA colour .................................................................................... 9610020001
- 6 Suction electrodes ......................................................................................................................... 9610054000
- ECG electrode cream (100 ml) ...................................................................................................... 99210091133
- Recording Paper, Z-folded (1 Pack) ................................................................................................ 9690029000
- Carrying Case ................................................................................................................................... 9100048000
- Recording Paper, Roll (Pack with 20 Rolls) ................................................................................... 9690029020
- Trolley .............................................................................................................................................. 9704250100
- Patient Cable Holder ........................................................................................................................ 9101341000
- User Guide (French / German / Italian) ........................................................................................... 9740440001
- User Guide (English / Spanish / Portuguese) .................................................................................. 9740440002

P80 Page 41
07.2001

Ordering Information

ENGLISH
Complete Units
P80 Basic, 230 V; Complete with standard accessories ....................................................................... 9840400010
P80 Measurements, 230 V; Complete with standard accessories ......................................................... 9840400020
Standard Accessories:
Patient cable EURO colours; pack of paper; set of ten electrodes
ECG gel; User Guide
P80 Basic, 115 V; Complete with standard accessories ....................................................................... 9840400011
P80 Measurement, 115 V; Complete with standard accessories .......................................................... 9840400021
Standard Accessories:
Patient cable USA colours; pack of paper; set of ten electrodes
ECG gel; User Guide

Accessories
- 10-lead Patient Cable, Euro ............................................................................................................. 9740410001
- 10-lead Patient Cable, USA ............................................................................................................. 9740410002
- 4 Extremity clamp-electrodes .......................................................................................................... 9610020000
- 4 Extremity clamp-electrodes, USA colour .................................................................................... 9610020001
- 6 Suction electrodes ......................................................................................................................... 9610054000
- ECG electrode cream (100 ml) ...................................................................................................... 99210091133
- Recording Paper, Z-folded (1 Pack) ................................................................................................ 9690029000
- Carrying Case ................................................................................................................................... 9100048000
- Recording Paper, Roll (Pack with 20 Rolls) ................................................................................... 9690029020
- Trolley .............................................................................................................................................. 9704250100
- Patient Cable Holder ........................................................................................................................ 9101341000
- User Guide (French / German / Italian) ........................................................................................... 9740440001
- User Guide (English / Spanish / Portuguese) .................................................................................. 9740440002

P80 Page 41
07.2001
Technical Data
Technical data subject to change without notice.
Dimensions (l/w/h): 296 x 220 x 72.5 mm
Weight: 2.9 kg
Mains Supply: 115 / 230 VAC, 50/60 Hz
Battery: Built-in 12 V lead-acid battery (rechargeable)
Power Consumption: Recording: 28 VA max.
Leads: Standard / Cabrera
Paper Speed: 5 / 25 / 50 mm/s (direct)
Sensitivity: 5 /10 / 20 mm/mV,
either automatically adjusted or manually selected
Chart Paper: Thermoreactive
- Z-folded, 90 mm wide, perforation 70 mm or
- Roll, 90 mm wide
Printing Process: High-resolution thermal print head, 8 dots per mm
Recording Tracks: 3 channels, positioned at optimal width on 80 mm, automatic baseline
adjustment
Automatic Lead Programs: 3 channel representation of
12 simultaneously acquired standard leads
Data Record: - Listing of ECG recording data
- Version M:
ECG measurement results (intervals, amplitudes, electrical axes),
average complexes with optional measurement reference markings.

Page 42 P80
07.2001

Technical Data
Technical data subject to change without notice.
Dimensions (l/w/h): 296 x 220 x 72.5 mm
Weight: 2.9 kg
Mains Supply: 115 / 230 VAC, 50/60 Hz
Battery: Built-in 12 V lead-acid battery (rechargeable)
Power Consumption: Recording: 28 VA max.
Leads: Standard / Cabrera
Paper Speed: 5 / 25 / 50 mm/s (direct)
Sensitivity: 5 /10 / 20 mm/mV,
either automatically adjusted or manually selected
Chart Paper: Thermoreactive
- Z-folded, 90 mm wide, perforation 70 mm or
- Roll, 90 mm wide
Printing Process: High-resolution thermal print head, 8 dots per mm
Recording Tracks: 3 channels, positioned at optimal width on 80 mm, automatic baseline
adjustment
Automatic Lead Programs: 3 channel representation of
12 simultaneously acquired standard leads
Data Record: - Listing of ECG recording data
- Version M:
ECG measurement results (intervals, amplitudes, electrical axes),
average complexes with optional measurement reference markings.

Page 42 P80
07.2001
Technical Data

ENGLISH
ECG Storage: - Memory for 10 s, 12-lead ECG
- Circular input memory for 10 s, 12-lead ECG.
Frequency Range
of Digital Recorder: - 0 to 150 Hz (IEC)
- 0 to 150 Hz (AHA)
ECG Amplifier: - Simultaneous, synchronous registration of all
9 active electrode signals (= 12 standard leads)
- Sampling frequency: 1000 Hz
- Digital resolution: 5 µV
- Dynamic range: ±9 mVAC
- Max. electrode potential: ±300 mVDC
- Time constant: >3.2 s
- Frequency response: 0.05 to 150 Hz (-3 dB)
- Input impedance: >10 MΩ
Myogram Filter
(muscle tremor filter): 25 Hz or 35 Hz, programmable (not active on averaged waveform). The
stored ECGs can be printed with or without filter.
Line Frequency Filter: Distortion-free suppression of superimposed 50 or 60 Hz sinusoidal in-
terferences by means of an adaptive digital filter.
Patient Input: Fully floating and isolated, defibrillation protected
Patient Leakage Current: <5 µA
Safety Standard: CF according to IEC 601-1 and IEC 601-2-25
Safety Class: I according to IEC 601-1

P80 Page 43
07.2001

Technical Data
ECG Storage: - Memory for 10 s, 12-lead ECG

ENGLISH
- Circular input memory for 10 s, 12-lead ECG.
Frequency Range
of Digital Recorder: - 0 to 150 Hz (IEC)
- 0 to 150 Hz (AHA)
ECG Amplifier: - Simultaneous, synchronous registration of all
9 active electrode signals (= 12 standard leads)
- Sampling frequency: 1000 Hz
- Digital resolution: 5 µV
- Dynamic range: ±9 mVAC
- Max. electrode potential: ±300 mVDC
- Time constant: >3.2 s
- Frequency response: 0.05 to 150 Hz (-3 dB)
- Input impedance: >10 MΩ
Myogram Filter
(muscle tremor filter): 25 Hz or 35 Hz, programmable (not active on averaged waveform). The
stored ECGs can be printed with or without filter.
Line Frequency Filter: Distortion-free suppression of superimposed 50 or 60 Hz sinusoidal in-
terferences by means of an adaptive digital filter.
Patient Input: Fully floating and isolated, defibrillation protected
Patient Leakage Current: <5 µA
Safety Standard: CF according to IEC 601-1 and IEC 601-2-25
Safety Class: I according to IEC 601-1

P80 Page 43
07.2001
Technical Data
Environmental Conditions: - Temperature, Operating: 10° to 40° C
- Temperature, Storage: -10° to 55° C
- Relative humidity: 25 to 95% (non condensing)
- Atmospheric pressure: 700 to 1060 hPa
Control Panel: Rubber keys
Technical data subject to change without notice.

Available Configurations
The P80 is available in two different versions:
Standard Version: Unit with ECG recording and printout capabilities.
Version M: Unit with additional ECG measurement program.

Page 44 P80
07.2001

Technical Data
Environmental Conditions: - Temperature, Operating: 10° to 40° C
- Temperature, Storage: -10° to 55° C
- Relative humidity: 25 to 95% (non condensing)
- Atmospheric pressure: 700 to 1060 hPa
Control Panel: Rubber keys
Technical data subject to change without notice.

Available Configurations
The P80 is available in two different versions:
Standard Version: Unit with ECG recording and printout capabilities.
Version M: Unit with additional ECG measurement program.

Page 44 P80