Arco Cirúrgico Ge Everview 7500 - Ms
Arco Cirúrgico Ge Everview 7500 - Ms
Arco Cirúrgico Ge Everview 7500 - Ms
2415170-100
Rev 5
© 2005, 2006, 2007, 2008
GE HUALUN Medical Systems
All Rights Reserved
Revision History
Date Rev Change Description
Aug. 2005 0 Initial Release
Nov. 2005 1 Add labels
Mar. 2006 2 Update due to design changes
Oct. 2006 3 Update due to design changes from M4 to M5
Jun. 2007 4 Update contents due to design changes
Jan. 2008 5 Update contents
IMPORTANT SAVE THESE INSTRUCTIONS. . PLEASE READ THIS MANUAL BEFORE USING EverView 7500 C-Arm X-Ray Imaging System.
This manual may not be reproduced, in whole or in part, without the written permission of GE Healthcare.
Other product and company names mentioned herein are the property of their respective owners.
The contents of this document are accurate at the time of publication. However, changes in design and additional features can, at
any time, be incorporated in the hardware and software and may not be reflected in this version of the document. Contact GE
Healthcare Technical Support for clarification, if discrepancies arise.
The text of this manual was originally written, approved and published by the manufacturer in English.
GE HUALUN Medical Systems, a General Electric company, going to market as GE Healthcare.
Page II
Introduction and Safety
警告 本维修手册仅提供英文版本。
(ZH-CN)
如果维修服务提供商需要非英文版本,客户需自行提供翻译服务。
未详细阅读和完全理解本维修手册之前,不得进行维修。
忽略本警告可能对维修人员,操作员或患者造成触电、机械伤害或其他形式的伤害。
VÝSTRAHA Tento provozní návod existuje pouze v anglickém jazyce.
(CS)
V případě, že externí služba zákazníkům potřebuje návod v jiném jazyce, je zajištění
překladu do odpovídajícího jazyka úkolem zákazníka.
Nesnažte se o údržbu tohoto zařízení, aniž byste si přečetli tento provozní návod a
pochopili jeho obsah.
Page III
EverView 7500 C-Arm X-Ray Imaging System
Failure to heed this warning may result in injury to the service provider, operator, or
patient from electric shock, or from mechanical or other hazards.
älä yritä korjata laitteistoa ennen kuin olet varmasti lukenut ja ymmärtänyt tämän
huolto-ohjeen.
Page IV
Introduction and Safety
ΜΗΝ ΕΠΙΧΕΙΡΗΣΕΤΕ ΤΗΝ ΕΚΤΕΛΕΣΗ ΕΡΓΑΣΙΩΝ ΣΕΡΒΙΣ ΣΤΟΝ ΕΞΟΠΛΙΣΜΟ ΕΚΤΟΣ ΕΑΝ ΕΧΕΤΕ
ΣΥΜΒΟΥΛΕΥΤΕΙ ΚΑΙ ΕΧΕΤΕ ΚΑΤΑΝΟΗΣΕΙ ΤΟ ΠΑΡΟΝ ΕΓΧΕΙΡΙ∆ΙΟ ΣΕΡΒΙΣ.
brot á sinna þessari aðvörun getur leitt til meiðsla á þjónustuveitanda, stjórnanda eða
sjúklings frá raflosti, vélrænu eða öðrum áhættum.
Page V
EverView 7500 C-Arm X-Ray Imaging System
IL NON RISPETTO della presente avvertenza potrebbe far compiere operazioni da cui derivino
lesioni all'addetto alla manutenzione, all'utilizzatore ed al paziente per folgorazione elettrica,
per urti meccanici od altri rischi.
このサービスマニュアルには英語版しかありません。
(JA)
サービスを担当される業者が英語以外の言語を要求される場合、翻訳作業はその業者
の責任で行うものとさせていただきます。
このサービスマニュアルを熟読し理解せずに、装置のサービスを行わないでくださ
い。
この警告に従わない場合、サービスを担当される方、操作員あるいは患者さんが、感
電や機械的又はその他の危険により負傷する可能性があります。
경고
본 서비스 지침서는 영어로만 이용하실 수 있습니다.
(KO)
Page VI
Introduction and Safety
Page VII
EverView 7500 C-Arm X-Ray Imaging System
NÃo tente reparar o equipamento sem ter consultado e compreendido este Manual de
AssistÉncia Técnica
O nÃo cumprimento deste aviso pode por em perigo a segurança do técnico, operador ou
paciente devido a choques elétricos, mecÂnicos ou outros.
Page VIII
Introduction and Safety
No se deberá dar servicio técnico al equipo,sin haber consultado y comprendido este manual
de servicio
La no observancia del presente aviso puede dar lugar a que el proveedor de servicios, el
operador o el paciente sufran lesiones provocadas por causas eléctricas, mecánicas o de
otra naturaleza.
FÖRSÖK INTE UTFÖRA SERVICE PÅ UTRUSTNINGEN OM DU INTE HAR LÄST OCH FÖRSTÅR DEN
HÄR SERVICEHANDBOKEN.
OM DU INTE TAR HÄNSYN TILL DEN HÄR VARNINGEN KAN DET RESULTERA I SKADOR PÅ
SERVICETEKNIKERN, OPERATÖREN ELLER PATIENTEN TILL FÖLJD AV ELEKTRISKA STÖTAR,
MEKANISKA FAROR ELLER ANDRA FAROR.
DİKKAT
BU SERVİS KILAVUZUNUN SADECE İNGİLİZCESİ MEVCUTTUR.
(TR)
EĞER MÜŞTERİ TEKNİSYENİ BU KILAVUZU İNGİLİZCE DIŞINDA BİR BAŞKA LİSANDAN TALEP
EDERSE, BUNU TERCÜME ETTİRMEK MÜŞTERİYE DÜŞER.
Page IX
EverView 7500 C-Arm X-Ray Imaging System
REGULATORY REQUIREMENTS
This product complies with the regulatory requirements of the following:
Council Directive 93/42/EEC concerning medical devices: the CE 0459
label affixed to the product testifies compliance to the Directive.
The location of the CE 0459 label is described in page 6-8 of Operator manual.
European registered place of business:
GE Medical System Europe
Quality Assurance and safety Regulatory Manager
BP34
Address: 283 Rue de la Miniere 78533 Buc Cedex France
Tel:+33 (0) 1 30 70 40 40
International Electrotechnical Commission (IEC), international standards organization, when applicable.
Page X
Introduction and Safety
Contents
Chapter 1 Introduction and Safety................................................................................................................... 1-1
1.1 Introduction .......................................................................................................................................1-2
1.1.1 Objective ..........................................................................................................................................1-2
1.1.2 Scope .................................................................................................................................................1-2
1.1.3 Service Philosophy.......................................................................................................................1-2
1.1.4 Qualifications.................................................................................................................................1-2
1.1.5 Unauthorized Modifications....................................................................................................1-2
1.1.6 Certified Components................................................................................................................1-2
1.1.7 Manual Organization and Use................................................................................................1-3
1.2 Safety....................................................................................................................................................1-3
1.2.1 Safety Hazard Alerts ...................................................................................................................1-3
1.2.2 Ingress of Fluids ............................................................................................................................1-4
1.2.3 LOTO procedures..........................................................................................................................1-4
1.2.4 Electrical Shock .............................................................................................................................1-8
1.2.5 Implosion..........................................................................................................................................1-9
1.2.6 Explosion ..........................................................................................................................................1-9
1.2.7 Thermal Harm................................................................................................................................1-9
1.2.8 Equipment Stability .....................................................................................................................1-9
1.2.9 X-Ray Radiation .........................................................................................................................1-10
1.2.10 Motorized Mechanical Movement ..................................................................................1-10
1.2.11 Safety Interlock........................................................................................................................1-10
1.2.12 Labels...........................................................................................................................................1-10
1.2.13 Symbols.......................................................................................................................................1-17
1.3 Electromagnetic Compatibility Statement........................................................................1-18
Chapter 2 Systems OverView.............................................................................................................................. 2-1
2.1 System Descriptions ......................................................................................................................2-2
2.2 Motions of the C-Arm.....................................................................................................................2-4
2.3 EverView 7500 C-Arm X-Ray Imaging System Block ......................................................2-5
2.4 EverView 7500 C-Arm X-Ray Imaging System Schematics .........................................2-6
Chapter 3 Installation ............................................................................................................................................. 3-1
3.1 Pre-Installation .................................................................................................................................3-2
3.1.1 Dimensions and Weights of System....................................................................................3-2
3.1.2 Site Requirement ..........................................................................................................................3-2
3.1.3 Power Supply Requirements ...................................................................................................3-2
3.1.4 Environments Requirements...................................................................................................3-3
3.2 Unpacking...........................................................................................................................................3-4
3.3 Power Supply Adaptation .........................................................................................................3-10
3.3.1 OverView.......................................................................................................................................3-10
3.3.2 Change the System Frequency........................................................................................... 3-10
3.3.3 Change the Power Supply Cables and plugs................................................................3-11
3.3.4 Voltage Adaption.......................................................................................................................3-14
3.4 Software Configuration .............................................................................................................3-15
3.5 Options Installation .....................................................................................................................3-16
3.5.1 Sony Video Printer Installation ............................................................................................3-16
3.5.2 CDAP InstallationC ....................................................................................................................3-23
3.5.3 Advanced Configuration Option Installation ................................................................3-32
3.6 Installation Check.........................................................................................................................3-36
Page XI
EverView 7500 C-Arm X-Ray Imaging System
Page XII
Introduction and Safety
Page XIII
EverView 7500 C-Arm X-Ray Imaging System
Page XIV
System Overview
Chapter 1
Introduction and
Safety
Page 1-1
EverView 7500 C-Arm X-Ray Imaging System
1.1 Introduction
1.1.1 Objective
Provide EverView 7500 C-Arm X-Ray Imaging System service documentation that is consistently
organized, easy to access and is designed to aggressively expedite fault isolation and decrease
product down-time.
1.1.2 Scope
This manual’s contents document the necessary information for service.
1.1.4 Qualifications
Reading this manual does not qualify an untrained person to service the equipment. This
publication assumes the reader is a skilled and fully qualified electronic technician who has
experience servicing medical imaging systems. Factory trained service personnel will benefit the
most from this material.
Page 1-2
Introduction and Safety
1.2 Safety
Do not attempt to service or maintain this equipment before reading this entire section.
CAUTION Caution indicates a potentially hazardous situation which, if not avoided, may
result in moderate to minor injury, equipment damage or loss of data.
Page 1-3
EverView 7500 C-Arm X-Ray Imaging System
The equipment is not rated for water-tight operation. Do not allow any liquids to
WARNING
leak into the equipment where they can cause electric shock, fires, and short
circuits. If leakage does occur, ensure that all internal circuitry is completely dry
before attempting to operate the equipment.
Never store or operate the system in the presence of conductive fluids like water
WARNING or saline solution unless the system is adequately protected by an approved
bagging or draping system.
WARNING For the EverView 7500 system service, the following procedures are
mandatory and must be implemented by “Authorized Employees.”
TYPE SOURCE
ELECTRICAL 220/230/240/110/120 VAC
Page 1-4
Introduction and Safety
Page 1-5
EverView 7500 C-Arm X-Ray Imaging System
13. Verify zero volts DC across Pin 7 and Pin 5 in MPJ1 socket of MP6 board.
Page 1-6
Introduction and Safety
Page 1-7
EverView 7500 C-Arm X-Ray Imaging System
The equipment must be serviced only by personnel familiar with proper safety
WARNING
procedures for working near these components.
Have someone watch while you work near high voltages. This person must
WARNING remain clear of all circuitry and be prepared to turn off the power to the system
and render aid in an emergency.
High voltage cables, electrolytic capacitors, and CRTs can retain a dangerous
WARNING static charge for long periods after power has been removed. Some devices can
acquire a charge spontaneously without direct contact with other circuitry. Do
not touch these components unless power has been completely removed and
they have been discharged.
Never operate the system unless all potential equalization conductors (green
CAUTION wire with a yellow stripe) and related components are fastened in place.
Dangerous electrical shock and improper equipment operation can otherwise
result.
Remove all metal rings and watchbands before working on system circuitry. Skin
CAUTION burns and damage from involuntary muscle contractions can result if metal
jewelry shorts electrical circuits.
Page 1-8
Introduction and Safety
1.2.5 Implosion
EverView 7500 C-Arm X-Ray Imaging System has a Cathode Ray Tube (CRT) , do not locate
objects so that they might fall and strike the tube causing it to implode. Use caution when
working around the tube. The coating on the glass can also produce toxic dust and fumes. If a
CRT implodes:
1. Remove power immediately and evacuate the area.
2. Contact site administration regarding hazardous material disposal.
1.2.6 Explosion
If an abnormal condition occurs, such as the room fills with flammable gas, steps must be taken
to prevent the gas from contacting components within the equipment. Follow these guidelines:
1. Do not turn the system on or off.
2. Do not un-plug or plug-in system or peripheral equipment power cords.
3. Do not operate light switches or other electrical equipment.
4. Evacuate all personnel immediately.
5. Notify the hospital administration and/or fire department.
6. Ventilate the room to clear the air of all flammable vapor or gas.
7. Identify and remove the source of the flammable liquid, vapor, or gas.
Page 1-9
EverView 7500 C-Arm X-Ray Imaging System
1.2.12 Labels
Even though this manual is supplied only in English, a foreign language equivalent
NOTE
of the following labels may appear on your system.
1. Quantity: 1
This label indicates the position of the C-Arm
Wig-Wag movement handle. When the handle is
rotated to vertical position, the Wig-Wag
brake is released. When the handle is in
horizontal position, the Wig-Wag brake is
locked.
2. Quantity: 1
This label indicates the position of the C-Arm
Wig-Wag movement handle. When the handle is
rotated to vertical position, the Wig-Wag
brake is released. When the handle is in
horizontal position, the Wig-Wag brake is
locked.
Page 1-10
Introduction and Safety
3. Quantity: 2
This label indicates the position of the C-Arm
orbital movement and rotation handle. When the
handle is rotated to the dashed position, the
brake is released. When the handle is in
horizontal position, the brake is locked.
4. Quantity: 2
This symbol indicates the position of the C-Arm
orbital movement and rotation handle. When the
handle is rotated to the dashed position, the
brake is released. When the handle is in
horizontal position, the brake is locked.
5. Quantity: 1
This symbol indicates the position of the C-Arm
horizontal movement handle. When the handle
is rotated to the horizontal position, the brake
is released. When the handle is in vertical
position, the brake is locked.
6. Quantity: 1
7. Quantity: 1
This label indicates that the Mobile C-Arm’s mechanical
subassemblies position should be placed in their most
compact position before transporting or moving the system.
Refer to OM chapter 4 Mechanical Movement for details.
Page 1-11
EverView 7500 C-Arm X-Ray Imaging System
8. Quantity: 1
Attention, see accompanying documentation. This
symbol on the tube housing means that you can
9. Quantity: 1
This label on the control panel of the C-Arm
means that you can refer to Section 2.4 in this
! manual for instructions of the control panel.
10. Quantity: 1
This label near the orbital brake handle means
that there is accessible moving parts and keep
hands/fingers away.
11. Quantity: 1
This label near the safety switch means that you
must take care to rotate the C-Arm larger than
! 155°.
12. Quantity: 1
This label on the Workstation means that you can
refer to the WorkStation keyboard section in this
! manual for detailed information.
13. Quantity: 1
This label on the USB interface means you should
open the USB cover then insert the USB stick.This
! interface is only used for data transmission,
not for control.
14. Quantity: 1
This label on the footswitch means that the
degree of protection of the footswitch is IP X8.
IPX8 The electrical mechanism within the footswitch is
protected from the effects of continuous
immersion in water.
Page 1-12
Introduction and Safety
15. Quantity: 1
This symbol indicates the location of Emergency
Stop Switch on the EverView 7500 equipment.
standards.
NOTE Reserve the share key carefully in case the system could not be reset.
16. Quantity: 1
This mark means that there is ionization radiation.
17. Quantity: 1
System nameplate.
Indicates manufacture information and input
power requirements.
This label is prefabricated on the system in Chinese, and the other version of the
NOTE labels are delivered with the system. It can be installed by GE qualified service
engineer according to the requirements of the local users.
18. Quantity: 1
This label indicates the system complies with the
US Federal Performance standard 21 CFR
Subchapter J and applicable international
standards.
This label is only on the system mounted with DAP unit (Option 2). It can be
NOTE
installed by GE qualified service engineer when the system is mounted with DAP
unit (Option 2).
19. Quantity: 1
X-ray Generator/Collimator
P.N.: 2397125
S.N.:
RATING VOLTAGE: 110kV
TUBE MODEL: 110/3 DF
INH. FILTERATION: 3.5mm AL
TYPE B
FOCAL SPOTS: 1.4mm 0.6mm
COLLIMATOR ASSEMBLY: 2415086
MANUFACTURED:
This equipment is certified to be in compliance with the applicable standards of the
Code of Federal Regulations,Title 21, Subchapter J, as of the date of manufacture.
GE HUALUN MEDICAL SYSTEMS CO. Ltd.
No 1 Yong Chang North Road Beijing Economic-Technological Development
Area 100176 BEIJING CHINA.
This symbol Indicates manufacture information
and input power requirements of X-Ray generator.
Page 1-13
EverView 7500 C-Arm X-Ray Imaging System
20. Quantity: 1
This symbol indicates the system was tested by
a Notified Body and was found to be in
0459 compliance with the requirements of all
relevant directives and standards in effect
within the European Union at the time of
manufacture.
21. Quantity: 1
This symbol indicates the generator was tested
by a Notified Body and was found to be in
0459 compliance with the requirements of all
relevant directives and standards in effect
within the European Union at the time of
manufacture.
22. Quantity: 1
2339039,CASSETTE HOLDER
This is the mark of the cassette holder.
23. Quantity: 1
24. Quantity: 2
This label indicates that waste electronic and
electrical equipment must not be disposed of as
unsorted municipal waste and must be collected
separately. Please contact an authorized
representative of the manufacturer for
information concerning the decommissioning of
your equipment.
25. Quantity: 1
This symbol indicates the system was tested by CCC
authorized body and was found to be in
compliance with the requirements of all
S
relevant directives and standards at the time
of manufacture.
Page 1-14
Introduction and Safety
26. Quantity: 1
This Label indicates the system was tested by ITS
authorized body and was found to be in
T ERT E
I N K
CM ETL CLASSIFIED
C US
Medical Equipment compliance with the requirements of all
relevant directives and standards at the time
L I
S T E D
9801745
of manufacture.
27. Quantity: 1
28. Quantity: 1
This is the mark of hard disk.
HARD DISK
P.N. 2414947
29. Quantity: 1
30. Quantity: 1
!
31. Quantity: 1
INDUCTANCE
This is the mark of inductance.
P.N. 2339247
32. Quantity: 1
INDUCTANCE
This is the mark of inductance.
P.N. 2339246
Page 1-15
EverView 7500 C-Arm X-Ray Imaging System
33. Quantity: 1
34. Quantity: 1
N L1 N1 L
This is the mark of J1 connector.
1 2 3 4 5 6 7 8 9 10 11 12
J1
2
3
6
7
8
1
35. Quantity: 1
This is the mark of J2 connector.
9
13
440v
14
10
11 15
Fusel 1
J2 220v 16
12
Fusel 1
17
13 Fusel 2
24v 18
Fusel 2
14
36. Quantity: 1
L N
This is the mark of monitor connector.
24V Indicator
220V N
220V L
37. Quantity: 2
Page 1-16
Introduction and Safety
1.2.13 Symbols
Be familiar with the following symbols that may appear on equipment and schematics, so you
can safely maintain and operate the system:
Dangerous Voltage
This symbol identifies areas where hazardous voltages may be present. Use
appropriate safety precautions.
X-Ray Source
This X-Ray Source symbol can be found on the X-Ray generator. It indicates the
location of the X-Ray source and the controls used to produce ionizing X-Radiation.
Use appropriate precautions and protective equipment when using the system
and at all times when X-Rays are present.
This symbol is used to identify controls that will produce ionizing X-Radiation
when activated. Always use appropriate precautions when generating
X-Rays.
This symbol identifies system terminals which, when connected together, are at the
same potential (not necessarily earth ground). Green/yellow wires normally connect
potential equalization terminals together.
Attention
Type B
This symbol indicates Type B applied parts. User is protected against electric shock by
protective earth grounding of all accessible conductive parts or the use of
non-conductive parts.
Page 1-17
EverView 7500 C-Arm X-Ray Imaging System
This equipment generates, uses, and can radiate radio frequency energy. The
NOTE
equipment may cause or be subject to radio frequency interference with other
medical and non–medical devices and radio communications. To provide
reasonable protection against such interference, the EverView 7500 system
complies with emissions limits for a Group 1, Class A Medical Devices and has
applicable immunity level as stated in EN 60601-1-2:2001.
Page 1-18
Introduction and Safety
NOTE To comply with the regulations applicable to an electromagn etic interface for a
Group 1, Class A Medical Device, and to minimize interference risks, the following
requirements shall be applied:
1. All interconnect cables to peripheral devices must be shielded and
properly grounded. The use of cables which are not properly shielded
and grounded may result in the equipment causing radio frequency
interference in violation of the European Union Medical Device
directive and FCC regulations.
Page 1-19
EverView 7500 C-Arm X-Ray Imaging System
The EverView 7500 system is suitable for use in the specified electromagnetic environment.
The purchaser or user of the EverView 7500 system should assure that it is used in an
electromagnetic environment as described below:
Emissions Test Compliance Electromagnetic Environment
The EverView 7500 system uses RF energy only for
its internal function. Therefore, its RF emissions
RF Emissions Group1
are very low and are not likely to cause any
CISPR11 interference in nearby electronic equipment.
RF Emissions Class A The EverView 7500 system is suitable for use in all
CISPR11 establishments other than domestic and those directly
connected to the public low-voltage power supply
Harmonic emissions Not
network that supplies buildings used for domestic
applicable purposes.
IEC 61000-3-2
Voltage fluctuations/ Not
flicker emissions applicable
IEC 61000-3-3
Page 1-20
Introduction and Safety
Page 1-21
EverView 7500 C-Arm X-Ray Imaging System
3,5
Radiated RF 3 V/m [E1=] 3 V/m d =[ ] P
3
B B
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and
reflection from structures, objects and people.
*Field strengths from fixed transmitters, such as base stations for cellular telephones and land mobile radios, amateur
radio, AM and FM radio broadcast, and TV broadcast cannot be estimated accurately. To assess the electromagnetic
environment due to fixed RF transmitters, an electromagnetic site survey should be performed. If the measured field
strength exceeds the RF compliance level above, observe the EverView 7500 system to verify normal operation in each
use location. If abnormal performance is observed, additional measures may be necessary, such as re-orienting or
relocating the [EQUIPMENT and/or SYSTEM].
Page 1-22
Introduction and Safety
**Over the frequency range 150 kHz to 80 MHz, field strengths are less than 3 V/m.
The Recommended Separation Distances are listed in the next table.
Note: These are guidelines. Actual conditions may vary.
Frequency of
Transmitter
150KHz to 80 MHz 80 MHz to 800 MHz 800 MHz to 2,5 GHz
Equation
3,5 3,5 7
d =[ ] P d =[ ] P d =[ ] P
3 3 3
Rated Power of
Transmitter DISTANCE DISTANCE DISTANCE
(W) (meters) (meters) (meters)
100 12 12 23
For transmitters rated at a power not listed above, the DISTANCE can be estimated using the equation
in the corresponding column, where P is the power rating of the transmitter in watts (W) according to
the transmitter manufacturer.
Page 1-23
System Overview
Chapter 2
Systems OverView
Page 2-1
EverView 7500 C-Arm X-Ray Imaging System
The EverView 7500 C-Arm X-Ray Imaging System is intended to provide fluoroscopic and film
imaging of the patient in diagnostic and surgical applications. Clinical applications may include,
but are not limited to, surgery, critical care and emergency room procedures. This product will
be used on a daily basis by such users. The product may be used for other imaging
applications at the physician’s discretion.
EverView 7500 C-Arm X-Ray imaging system includes:
• Mobil C-Arm
Is composed of generator, monoblock, collimator, image intensifier, CCD camera head
and control panel.
• Mobil WorkStation
Is composed of 2 monitors, 1 computer hard disk, keyboard, etc.
• Options
Sony Thermal Printer
DAP
Cassette holder, 10×12” (24cm×30 cm)
Spacer (30 cm)
Page 2-2
System Overview
The C-Arms derive their name from the rigid, C-shaped weldment that supports the Image
Intensifier, collimator, CCD Camera, and X-ray tube. These components together are commonly
called the “C”, and the entire assembly from the casters up is known as the “C-Arm”.
The X-Ray tube generates an X-ray beam that is shaped and oriented by the collimator. After
passing through patient anatomy, the beam is directed onto a film cassette to produce a
radiographic film image, or through the image intensifier and CCD camera, to convert the X-ray
image to a video signal that is transmitted to the WorkStation for display, processing, and
storage.
Page 2-3
EverView 7500 C-Arm X-Ray Imaging System
Orbital Movement
Page 2-4
System Overview
Collimator
Filament PCB KV/MA PCB Exposure
Current Control KV & MA Control
Brightness Self-regulation
Management
of Filament
Tabletop MP4 MP3 Controller
MP2
Grid Driver PCB
MP1
II/CCD IBS
Operator Desk
CCU Display PCB Exp Foot Switch
Optics
50Hz CCIR X-Ray ON Warning Light
100Hz
Rotation
Sensor
Imaging Computer
75Hz VGA
CCD Recordt
17"
Monitor 2
752X582 17" Display
CCD Database
Monitor 1
Camera
Keyboard
Video
PAL
Workstation Out
Page 2-5
EverView 7500 C-Arm X-Ray Imaging System
PJ3-1
J1 Brn
1 BT1
9 Brn M PJ1-3 Brn
239071
A
1 13 4
A PJ3-3
2 Blk
Wh 2 440V 14 M PJ1-1 Blu 233945
8
3
t
Re 3 10
4 1 Gry J3B J5 J2 J3
J3A
PJ3-4
d 4
1
Org 15 Fuse1 2415974
J0 J5 241433
2339471
5 Yel 5 1220V Blk 1 5
6 Gry 12
Yel 16
Fuse1
2415973 J2
J4
CCU 5
2390715
PJ3-6 Blu 31 1
14 24V Blu
7 CJ4
CJ3
6 Display Board:
17
Fuse2 M X8-7
J1
8 7 51 22V CJ CJ 241433
3 2414090
Brn
CJ
M X8-6
Grn 7 CCD
6
9 8 6 Fuse2 8 4
2415971 PJ1-1 17 Re 18
Connect Board :
10 2415957
PJ1-3 d
11
PJ1-5 Blu 2415950
2414092
Blu 12 241436
2415964 PJ1-9 9
For detail connection 2415955
please refer to page 3 I.I
J2-12
J2-13
AD1
AL3
AL4
PJ4-1 1 BT2
2415969
5 A:
MX3-3/4
M X2 1234567
MX1
M X1
MX3-5
PJ4-2 2
3
PJ4-3 3 M X1
MX8 AC AR
6 4
PJ4-4 4 AL3 AL4 AL5 AL6
PJ4-7
7 2415982 M P1 M P2 MP3 M P4 M P5 M P6 M P7
2415975
PJ4-8
1
M PJ1-
2415977 7
MX3
MP2-
MX15-
MX15-
M PJ1-
J1
Blk
B B
1
5
Red
Re FJ1-1 Re M PJ1-8 AD 1 AD2
d d 2 1
Blu FJ1-2 Blu Blu
C J8-4 2416039 I:
IJ
2415968
AL2-2
Blk AL2
CJ8-1 2415976
MPJ1
AL1-2
AL2-1
Em ergency
220VL
220VN
220VL
J2-13
220V
+24V
CO
GN
-15V
CO
CO
+5V
Sw itch
+15
M
S:
GN
M
V1
V1
M
D
N
J2-14
SJ
M PJ1-6 1
MX4
Brn
Mother Board ASM : 2
2415963 AL1
2415948 2415978 2415166 M X3-1/2
GND
Blu AL2-2
Red AL2-1
Brn AL1-2
2 J2-2 G ND Brn Blu
Blu
Brn
PJ7
3 J2-3 Blk
PJ6
4 J2-4 2339469
5 J2-5
H:
MX5-5Red
MX5-4Grn
2416052
PDU Board: 233947 H J1
PJ8
PJ5-1
Wht
MX5-
0
2414086
PJ5
PJ5-4
6
241448 HJ2
2
2415980
CJ7-1
L
P
1
CJ7-3
2415979
PJ1
MD
PJ1-6 OX1 2415953
C I1 1 2 3 4 5
N C
PJ1-8 A
2372499
B
2372500 1 4
F:
C M otor
J3 100Hz FJ3
Video
+ Driving
Brn
Blu
J1-1
GND O: O X2 1 M X5-1
M X5-
J1-2
M ale 2 J1-3
J2 5 2 2PIN
50Hz Video 3
MX5- Fem alJ5-F 3 1
3
Filter GN D 4 B lue& Green e
M ale
FJ1 FJ2
2415947
16AW G
2415970 1 J5-M 2 1
Note! 2415949 2415951
Do the right wire
O X3
connection ? ? ? ? ? ? ? ? ? ? ? ? ? ? ?
2415981
2415967
2415965
betw een
the coxial cabel J4-M Male
Brn Blu and BN C J4-F Femal
3PIN
FT ? ? ? ? ? ? ? ? ? ? ? ?
2415972 connector 1 Proximity
Ou
L+
L-
2415966 Sensor
NP
a
a
a
Sensor ? ? /? ? ? ? /? ? ? ? /? ? ? ?
Breaker Breaker N
EverView 7500 II
Unlock
Unlock
K1 K2
1
W iring Diagram D
A
D
A
A
A
2415945
F.S
FT
? ? ? ? ? ? ? ? ? ? ?
S10002P
FT-EN
Brn
N Blu please refer to page 3
For detail connection 1
L 2
3 4
5
Double grounding ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? ? SCH 1
2339441(2415147)
IX2
connection
w ith tie-in protected by GE HUALUN MEDICAL SYSTEMS E 0
Fem al
e
heat-shrinking pipe 3
1 2 3 4 5 6
Page 2-6
System Overview
VGA
Key board
video in
power
241 47 25
printer
YJ-2 YJ-1 24 14 72 2
VGA
24 1472 4
power
24 1472 5
24 1472 5
female male
key board&mouse
4 3 2 1 4 3 2 1 video in
2
AWG18 Note: a grounding label should 24 14 72 6 3
Length =10cm be attached around here
4
GND
5
7
Shafler
24 1472 3
24 1472 4
Filter
J3 J2
OMRON
G2R-2-S-12VDC
1
5 4
6 3
7 2
8
220VN
220VL
GND
1 2 3 4 5
A
IX1 B
C
video
GND
GND
video
Page 2-7
Chapter 3
Installation
EverView 7500 C-Arm X-Ray Imaging system
3.1 Pre-Installation
3.1.1 Dimensions and Weights of System
Dimensions:
C-Arm: 1730mm x 760mm x 1700mm (Length x Width x Height )
WorkStation: 840mm x 650mm x 1630mm (Width x Depth x Height )
Weight:s:
C-Arm: 480 Kg (with packing) 250Kg (net weight)
WorkStation: 225 Kg (with packing) 150Kg (net weight)
The floor should be hard enough to support the system. It should also be
WARNING
smooth and horizontal.
Page 3-2
Installation
Page 3-3
EverView 7500 C-Arm X-Ray Imaging system
3.2 Unpacking
There are 2 packages for EverView 7500 C-Arm X-Ray Imaging system. One is for
NOTE C-Arm (L X W X H=1900mmX900mmX1970mm), the other is WorkStation (L X W X
H=800mmX960mmX1820mm). It is easy for one FE to do unpackage in 2 hours.
The WorkStation and Mobile C-Arm ramp is located in rear panel of the Mobile
NOTE
C-Arm’s enclosure. You should open the Mobile C-Arm’s enclosure to get the ramp
first.
Before unpacking, inspect the shipping enclosure /boxes for damage and report to
NOTE the shipper if available.
The unpacking of the components of the system before it is moved into the operation room will
be made for checking if there is damage or non-apparent damage occurring during
transportation.
The packages will be minutely examined during installation, in order not to overlook parts of
small dimensions or envelopes containing screws, washers, etc.
If damage is observed, immediately notify the transport service and the transport insurance
service of the factory.
The same will apply for non–apparent damage due to transport discovered when unpacking the
material or during installation, UP TO 15 DAYS MAXIMUM AFTER DELIVERY.
A transport service will generally not pay a bill for non-apparent damage unless an inspection
has been demanded within 15 days after delivery.
If damage is discovered, immediately contact the services concerned, specifying the type of
apparatus, the serial number and the order number if possible, and describing the nature of the
damage.
Transport the equipment to the operating site, check the parts according to packing list (refer to
the packing list with the system for details).
Page 3-4
Installation
Please follow the steps below to unpack the EverView 7500 C-Arm X-Ray Imaging System:
1. First , place the C-Arm package as pictures below and leave enough space for backward
moving of the C-Arm.
Front (Image
Intensifier) side of
the C-Arm, with a
block outside
Right panel
Rear panel
(ramp attached)
2. Unscrew the screws of the fastening angle iron. Remove the angle iron. Remove the top
cover first.
3. Remove the wood block on right side ( from rear side to front side view) of the package of
the C-Arm. Remove the right side panel. Remove the 2 crossbeams on the left side panel
as crowbars. Then remove the left panel and rear panel, front panel of the package.
Remove the ramp from rear panel and put it on the rear side of the C-Arm. Remove the
plastic on the C-Arm.
Page 3-5
EverView 7500 C-Arm X-Ray Imaging system
crossbeams
Wood block
4. Unscrew the 2 bolts with a 8 mm sleeve wrench and remove the support woods.
2 bolts
5. Loosen the fastening nut of the support assembly, then unscrew the screw bolt.
Page 3-6
Installation
6. Unscrew the 4 screws of the fastening assembly at the front of the C-Arm. Remove the
fastening assemblies with a crowbar.
7. Unscrew the 8 screws of the 2 fastening assemblies on each side of the C-Arm and
remove the fastening assemblies with a crowbar ( only right side for illustration).
8. Check and Verify that all the mechanical movements brakes are locked. Release the
wheel brake. Hold the right steering handle in a vertical position to keep the rear wheels
in a straight direction. Then carefully move the C-Arm backward along the ramp to the
site you desire.
The C-Arm weights about 250 Kg. Two persons should move it slowly to prevent
WARNING lose of control.
Page 3-7
EverView 7500 C-Arm X-Ray Imaging system
9. Remove the top cover and side covers of the WorkStation Package. Remove the front and
back fastening wood blocks.
10. Lift the WorkStation with a crowbar, and draw out the stow-woods under the
WorkStation. Put the wood ramp in position, and push the WorkStation along the ramp.
11. Move the WorkStation to the site near the C-Arm and lock the wheels.
12. Insert the male socket of the WorkStation’s integrated cable into the female socket
located on the right side of the C-Arm’s cover. Push the socket in and lock it in position.
13. Plug the C-Arm power plug into a properly rated AC receptacle. Power on the system.
Before power on the system, please check and make sure that the power
WARNING frequency and voltage of the system are compatible with the power supply. Or
you must change the power supply and frequency settings of the system.
Refer to Section 3.3 for details.
14. Press the Raise C-Arm Command Key to lift the C-Arm for about 20 cm height.
Page 3-8
Installation
15. Use a pliers or 17mm wrench to loosen the fasten nut then unscrew the screw rod.
Remove the backward-forward fastening assembly.
16. Loosen the horizontal movement brake handle and push the C-Arm forward. Lock the
brake handle. Lower the C-Arm to it’s lowest position and then power off the system.
17. Move it to the site you desire.
Page 3-9
EverView 7500 C-Arm X-Ray Imaging system
If your power is 60 Hz, please change the jumper on MP5 from 50 Hz to 60 Hz. The location of
the jumper is as the picture below.
Page 3-10
Installation
1. Open the cable kits: 110 V power cable kit (2415136).Remove the back cover of the
C-Arm control panel. Loosen the fasten screws of the incoming circuit breaker and
remove the 220 V power supply cable.
2. There are 2 types of plugs provided by GE Healthcare: European type and American type.
Select or buy the relative plugs according to your local regulatory requirements and
sockets style. They must be reliably grounded.
3. If the system is used in Canada and Australian and USA, please select and buy the
relative plugs according to the list below. The selected plugs shall be listed or recognized
type plugs. The manufacture doesn’t provide these types of plugs.
Countries Rated Current Comments Voltage Frequency
Canada 20 A Hospital Grade 120 V 60 Hz
Plug
Australian 16 A transparent 230 V 50 Hz
USA 20 A "Hospital grade" 120 V 60 Hz
or "Hospital
only" type plug
NOTE
Such plug shall be installed by GE service representative.
4. Connect the plug to the power supply cable and connect the power supply cable to the
incoming terminals of the K2 circuit breaker. Then change the bridge connectors of J1
terminal blocks and adjust the jumper on PDU board at the same time. ( The follows are
the schematics for 110 V and 220 V system FYR).
Page 3-11
EverView 7500 C-Arm X-Ray Imaging system
Jumper
Jumper
NOTE
You should change the 2 spacers’ position too.
Page 3-12
Installation
PDU board
Brn 9
PJ1-1
a
a
a
10
Breaker Breaker 110V
Unlock
Unlock
11
PJ1-3
K1 K2
Blu PJ1-5
A
A
A
A
12
PJ1-9 220V
Brn
L N Blu
PDU board
Brn
a
a
a
Unlock
Unlock
10
K1 K2
11
PJ1-3
A
A
A
A
PJ1-5
Blu 12
220V
PJ1-9
Brn
L N Blu
5. Plug the power supply plug into the socket and check if the system can power on
normally.
Page 3-13
EverView 7500 C-Arm X-Ray Imaging system
N L1 N1 L
J1
11
10
12
9
3
8
7
2
5
4
1
6
1
8
2
7
6
5
Page 3-14
Installation
Page 3-15
EverView 7500 C-Arm X-Ray Imaging system
CAUTION The Video Printer is not fixed in the cabinet. Please pay attention when moving the
WorkStation.
1. Open the rear cover of the Workstation (refer to section 3.6.2).
2. Plug one end of the video cable to the video port of the printer. Plug the power
supply cable into the socket on the printer. Put the cables through the hole of the
Workstation. Put the printer in the cabinet as pictures below.
3. Unscrew the screws of the draw-out hard disk. Plug the cable and video cable
connecter into the HUB indicated in the picture below.
Page 3-16
Installation
4. Cut off the other end of the power supply cable, strip off 3 wire ends, connect the 3
wires to the terminal blocks at the bottom of the workstation (brown line to the brown
power supply line, and blue to the blue, green-yellow to the green-yellow). Put the
power supply cable along the wiring harness to get a good layout.
NOTE After the printer is installed, put the micro-switch to the Hi position.
Page 3-17
EverView 7500 C-Arm X-Ray Imaging system
Page 3-18
Installation
2. Double-click [System]
The [System Properties] dialog box opens.
3. Click the [Hardware] tab.
4. Click the[Driver Signing ] button.
All the explanations in this guide assume that the setting in this dialog box is [Warn].
Installing the Printer Driver
1. Connect the UP-D897 to the computer.
2. Turn on the UP-D897.
3. Turn on the computer, exit the EverView 7500 application, refer to section 5.2.2 for details.
The [Found New Hardware Wizard] dialog box opens.
If the same [Found New Hardware Wizard] dialog box as in step 3 appears,
proceed to step 4.
If a [Found New Hardware Wizard] dialog box other than the one in step 3
appears ,proceed to step 5.
If the [Found New Hardware Wizard] dialog box does not appear, make sure the
UP-D897 is connected to the computer and turned on ,and then perform the
operations below. (If other USB printers are connected, disconnect all but the
UP-D897 before performing the following operations.)
Page 3-19
EverView 7500 C-Arm X-Ray Imaging system
5. Select [Install from a list or specific location(Advanced), and then click the [Next]
button.
Page 3-20
Installation
6. Select [Search for the best driver in these locations],select [Search removable media
(floppy ,CD-ROM)] ,and then insert the provided “PRINTER DRIVER & INSTRUCTIONS
FOR USE UP-D897”. CD-ROM into the CD-ROM drive.
7. Click the [Next] button.
The following dialog box opens.
Page 3-21
EverView 7500 C-Arm X-Ray Imaging system
After installation is finished, power on and check whether the printer can work normally.
Page 3-22
Installation
Electrical circuits inside the equipment use voltages that are capable of causing
WARNING serious injury or death from electrical shock, and/or damage to the equipment. Use
appropriate precautions.
To prevent ESD damage, ensure that you are properly grounded with a grounded
CAUTION
wrist strap before handling any electronic components sensitive to ESD.
NOTE The system being upgraded may differ slightly from what is illustrated.
1. Power off the EverView 7500 system and unplug the power supply plug of the system from
the socket.
2. Disconnect the integrated cable.
3. Unscrew the 4 screws of collimator cover and remove the collimator cover.
4. Insert the RJ45 plug of the 2 way power supply cable in the socket of the DAP kit.
Socket
RJ45 plug
Page 3-23
EverView 7500 C-Arm X-Ray Imaging system
The 8 way cable for DAP unit 2420029(Option 2) is different from 2 way cable for
NOTE
DAP unit 2413654(Option 1). Following pictures indicates difference.
6. Check the C-Arm and confirm the updating. Then power on the system, check if the display
and operation of the DAP work properly.
Page 3-24
Installation
Electrical circuits inside the equipment use voltages that are capable of causing
WARNING serious injury or death from electrical shock, and/or damage to the equipment. Use
appropriate precautions.
To prevent ESD damage, ensure that you are properly grounded with a grounded
CAUTION
wrist strap before handling any electronic components sensitive to ESD.
NOTE The system being upgraded may differ slightly from what is illustrated.
1. Power off the EverView 7500 system and unplug the power supply plug of the system
from the socket.
2. Disconnect the integrated cable.
3. Unscrew the 4 screws of tube cover and remove the tube cover.
Page 3-25
EverView 7500 C-Arm X-Ray Imaging system
Chamber
6. Place the DAP display module on the C-Arm consol and tighten the 2 screws as
indicated in following picture.
Module in chamber
7. Use socket to connect the RJ45 plug of the 8 way cable with the chamber cable plug.
NOTE
The 8 way cable for DAP unit 2420029(Option 2) is different from 2 way cable for
DAP unit 2413654(Option 1). Distinction is described above.
Socket
Plug of 8 way cable for DAP
unit 2420029(Option 2)
Page 3-26
Installation
8. Place the DAP display module on the C-Arm console and tighten the 2 screws as
indicated in following picture.
Page 3-27
EverView 7500 C-Arm X-Ray Imaging system
11.Use one end of the power cable of DAP kit to connect the power wire from the DAP
display module cable, as indicated in following picture.
Transfer socket
of DAP display
module
Power supply
Plug
12.At the other end of the power cable, there are red and black wires. Connect the red wire
in the power cable of DAP display module to +15V of power supply, and black wire to COM
of power supply.
NOTE Please test red, black terminal voltage and ensure the power source is +15 VDC.
Page 3-28
Installation
14.After all cables are connected, superpose the circular adjustment tool on the chamber
exactly, as indicated in following picture. The four longitudinal locking screws and
transverse locking screws are used to adjust chamber’s precision (detailed steps are
described below). Before doing adjustment of one direction, please loose the related
locking screws.
Transverse
locking
screws
Circular
adjustment tool
DAP chamber
Longitudinal
locking screws
Page 3-29
EverView 7500 C-Arm X-Ray Imaging system
Transverse
adjustment screw
Circular
adjustment tool
DAP chamber
Longitudinal
adjustment screw
Circular adjustment
tool image
Page 3-30
Installation
b) Fine adjustment: (Ensure the collimator work normally before this step)
Power on system, invoke HLF mode and set a high KV (about 70KV). Adjust collimator size
withing field size first, and then adjust the value of KV to make sure both the collimator's
edge and the circular adjustment tool' edge can be identified clearly one the screen.
Adjust the chamber longitudinal and transverse adjustment screws carefully to superpose
the two centers of circular adjustment tool and collimator. Decrease the size of collimator as
the picture shown, the exposure image center of circular adjustment tool and the collimator
should be coincident (No more than 2mm).
Circular adjustment
tool edge
Collimator edge
16.Remove the circular adjustment tool; fasten the eight locking screws tightly.
17.Check if DAP and DOSE function are normal under each filed size according to the
operating manual.
18.Fasten the cables with enlace strap to the side of collimator, reassemble the new tube
cover.
Page 3-31
EverView 7500 C-Arm X-Ray Imaging system
Enlace strap
USB port
Page 3-32
Installation
5. Attach the USB Dongle to the other end of the USB cable.
6. Fix the USB Dongle by binding the end of USB cable near the USB Dongle to the Mouse
and Keyboard cables and the carrying ring of the Dog to the upper cover of workstation
PC with the other 2 pieces of Enlace strap (refer to picture below).
Ring of the
USB Dongle
Enlace strap
binding
Page 3-33
EverView 7500 C-Arm X-Ray Imaging system
Option Inspection
When option is installed, perform the following operations to check the optional functionality.
Archive Device button should be enabled (see figure 3). If not, there is something wrong with the
option, and please use the service tool program to analyze and solve the problem.
4. Click Archive Device button to display Archive Device dialog to select to copy from system
to USB.
5. Plug a U-disk into the USB port.
6. Select some sample images, and click Copy.
The images should be properly copied from the system to the U-disk.
Page 3-34
Installation
The first, “ERROR. No dog is available.” which reveals that option is not yet installed or the dog
may get loose.
The second, “ERROR. The dog is corrupted.” This implies that the dog has been corrupted and it
should be replaced by a new one for the option to work properly.
The third, "OK. The dog works well." This is the result we prefer, which means the option is fine
and we need to do nothing with it.
Page 3-35
EverView 7500 C-Arm X-Ray Imaging system
Page 3-36
Installation
Transportation
Warning Label
The warning label (23) is on the control panel of the C-Arm. This label is installed on the system
only for USA market by local service engineer. Refer to the picture below for the location of the
label.
Page 3-37
EverView 7500 C-Arm X-Ray Imaging system
Control
Panel
Steering
Handles
Base &
Frame
Front and
back
wheels
Monoblock
Comments:
Indicate the location of damage of defects, if any, on the figure and add comments to describe
the damage if necessary.(Reference Mobile C-Arm Damage Inspection on previous page)
Page 3-38
Installation
Page 3-39
EverView 7500 C-Arm X-Ray Imaging system
Comments:
Indicate the location of damage or defects, if any, on the figure and add comments to describe
the damage if necessary. (Refer to WorkStation Damage Insepection on previous page.)
Page 3-40
Installation
Page 3-41
EverView 7500 C-Arm X-Ray Imaging system
Page 3-42
Installation
3. Remove the 8 screws of the WorkStation cover , remove the cover. Inspect the internal
cables. Check if there are loosen parts.
Page 3-43
EverView 7500 C-Arm X-Ray Imaging system
2. Remove the C-Arm’s covers and check the internal cables and PCB. Verify that there are no
loose parts. Follow the steps below to remove the covers.
a) Unsrcew the 2 screws of the strips(1 on each side of the cabinet), remove the
strips.
b) Unscrew the 8 fixing screws of the covers on each side of the cabinet.
Page 3-44
Installation
d) Remove the covers of the cabinet( covers arranged for illustration purpose).
Page 3-45
EverView 7500 C-Arm X-Ray Imaging system
• Brakes
1. Check the Wig-Wag brake operation. Check Wig-Wag for ease of movement (side to
side swing ) without excessive play when the brake handle is turned to release
position. When the brake handle is in lock position, check that it is hard to enable
Wig-Wag movement.
2. Check the Horizontal Movement brake operation. Extend and retract the Horizontal
Cross-Arm. Check for ease of movement without excessive play. When the brake
handle is in lock position, check that it is hard to enable Horizontal movement.
3. Check the rotation brake operation. Check the rotation for ease of movement without
excessive play. When the brake handle is in lock position, check that it is hard to
enable rotation movement.
4. Check the C-Arm orbital rotation brake operation. Check orbital rotation of the C-Arm
for ease of movement without excessive play. When the brake handle is in lock
position, check that it is hard to enable orbital rotation.
5. Check the operation of the rear wheel pedal brakes.
6. Check for proper operation of the steering handle and control of the rear wheels.
Check for ease of movement without excessive play.
7. Check the operation of the brakes of workstation.
• Steering
With the Mobile C-Arm’s rear wheels brake released, turn the right steering handle
90° to the left and right. Verify that the rear wheels turn 90° to the left and right
smoothly.
Place the Mobile C-Arm’s rear wheel brake in the released position. Push the
C-Arm slowly and verify that when the steering handle is turned the Mobile C-Arm
turns.
• Wheels
Push the C-Arm and verify that the wheels roll smoothly, the wheels turn freely.
Push the WorkStation and verify that the wheels roll smoothly without vibration and
that they are securely attached to the wheels.
Page 3-46
Installation
5. Press (Raise C-arm Command) key on the panel to raise the C-am, the
lower the C-arm. When the movement stops, press two keys at the same
time and the C-arm will be lowered to it’s lowest position.
Page 3-47
EverView 7500 C-Arm X-Ray Imaging system
WARNING
The system will produce X-Rays during the following procedures. Take
appropriate measures and precautions.
1. Manual Fluoroscopy
Place spacial resolution card in front of the image intensifier, the angle between the horizontal
axis of the image and the test card must be 45°。Select Manual Fluoroscopy mode
and appropriate kV value,take an exposure ,and stop when image is stable. The image is frozen
on the main monitor, verify the resolution in lp/mm is equal to, or greater than, the values listed
below. Repeat and select any of three field sizes (NORM, MAG1 and MAG2) to verify the resolution.
Field Size NORM MAG1 MAG2
Resolution (lp/mm) 1.4 1.8 2.2
2. Auto Fluoroscopy
Select Auto Fluoroscopy mode, and the NORM field size to take an exposure, verify that
the KV changes accurately as copper filters are inserted into the field in the following ranges:
Test Equipment : 1 mm Copper filter ( 3 for each system) No: P/N 00-887682-01.
(This equipment is not provided with the system by the manufacturer).
Page 3-48
Installation
Select Manual HLF mode + Pulse mode via the C-arm control panel, the
X-Ray Exposure indicator flashes when taking an exposure.
4. Manual HLF
Select Manual HLF mode + , verify that a short-term HLF exposure is taken
and automatically stopped; Audible tone beeps by the buzzer shows it’s in correct function.
5. Auto HLF
Place an object in front of the intensifier, select Auto HLF mode + to take
exposure. It will stop automatically.
6. I.I inspection
Place something on the I.I., take exposure under 3 fileds (9”, 6”,4.5”). Verify that the image
changes with the field.
7. Recursive Factor
rotation key (left then right), the image on the monitor will rotate relatively.
Enable the Motion Detection key, select Fluoroscopy mode , take an continuous exposure and
press the CCD rotation key , the image lag on the monitor will be smaller
than before.
Page 3-49
EverView 7500 C-Arm X-Ray Imaging system
9. Semi-transparent Shutters
Select manual fluoroscopy mode to take exposure intermittently. Open, close and rotate the
shutters to verify the functions. When the shutters are opened to the largest, the shutter will not
display on the monitor.
10. IRIS
Select manual fluoroscopy mode to take exposure intermittently, open and close the IRIS to verify
the functions. When the IRIS are opened to the largest, the IRIS will not display on the monitor.
Page 3-50
Installation
11. Radiography
Fasten the cassette holder on the intensifier; insert a cassette (10”×12”,with a film). Select
Radiography mode and take a short exposure under 50KV, 4mAs, verify that beeps are heard and
no errors are displayed, the developed film is clear.
Page 3-51
Chapter 4
Calibration
EverView 7500 C-Arm X-Ray Imaging system
4.1 Overview
The Calibration described herein should be performed each time you replace a major system
component. When possible, run these tests in sequence that appear in this section.
Tools
In addition to common tools, the following tools are recommended:
The system will produce X-Rays when implementing the following procedures.
WARNING Take appropriate measures and precautions to protect yourself and others from
X-Ray exposure.
You must obey the following steps to guarantee the safe operation of the C-Arm.
CAUTION Improper adjustment may cause premature failure of major system components.
Page 4-2
Calibration
MP3
Board
W3 jumper
Remove W Jumper,
Connect both probes
to W test point.
Page 4-3
EverView 7500 C-Arm X-Ray Imaging system
4.2.1 Measurement
1. Normal Fluoroscopy current measurement
Insert the MP3 PCB on one extended PCB and connect a DC current meter into the line of
MAMEAS2. (Remove the W jumper and connect the DC current meter probes to the two
pins).
2. High Level Fluoroscopy current measurement
Connect jumper HIDO on MP2 to disable automatic stop exposure function, please note
that exposure time should be as short as possible.
MP2 board
HIDO Jumper
Page 4-4
Calibration
4.2.2 Adjustment
Fluoroscopy current adjustment
1. Pre-adjust
a) Power on the system.
b) Select manual fluoroscopy mode.
c) Set the KV value to 110KV and measure the test point TP5 on MP3. The measured
value should be 9.3V (adjust POT3 on MP3 board to reach this value).
d) Set the KV value to 40KV and measure the test point TP5 on MP3. The measured
value should be about 0.3V (adjust POT3 on MP3 board to reach this value).
e) Measure the filament feedback on the resistance R51 on the MP4 board via
oscilloscope, take a 50kV exposure, adjust POT 7 on MP3 board, make the voltage
wave monitored by oscilloscope at the same level when exposure and after
exposure.
Page 4-5
EverView 7500 C-Arm X-Ray Imaging system
3. Take exposures. Check the current (0.5~4.0 mA) measured by the meter according to
the above values. The error should be within the ±10% ranges of the settings.
Otherwise adjust POT4 on MP4 board. Rotating it clockwise can increase the current ,
rotating it counter-clockwise can decrease the current.
4. To adjust 0.1mA accuracy, adjust POT 8 on MP3. POT 8 is only for low mA adjustment.
Raise kV step by step, otherwise instant high current may cause damage to
WARNING the equipment. 1/8 circle of knob in the meter will be a proper stroke to
prevent an instant high current.
Do not adjust the POT4 on MP4 board when unnecessary. Once do it, you
WARNING
must calibrate the HLF current and Radiography current (mAs).
5. Power off, disconnect extended board, put the jumper W back to extended board.
HLF current adjustment
HLF current adjustment can only be performed after Normal Fluoroscopy
NOTE
current adjustment.
1. Adjust
a. Put MP3 board on the extended board.
b. Set DC current meter to mA range.
c. Connect both probes to ‘W’ test point on the extended board
d. Remove ‘W’ jumper.
e. Connect jumper HIDO on MP2 board
f. Power on the system
g. Select manual HLF mode.
h. Set mA values as below:
mA 0.2 1.0 2.0 4.0 8.0
Measurement
2. Take exposure. Check the measured values and verify that the errors are within the ±10%
range of the values above. Otherwise, adjust POT 6 on MP3.
3. Power off the system.
4. Disconnect extended board, remove probes from W test point on extender board.
5. Disconnect jumper HIDO on MP2 board.
Page 4-6
Calibration
Page 4-7
EverView 7500 C-Arm X-Ray Imaging system
2. Timer adjust.
a) Set the following kV and mAs values:
1mAs(110kV) 80mAs(40kV) 4mAs(80KV) 20mAs (100KV)
b) Take exposure, check and verify the errors of the mAs value is within ±10% range.
c) If not, adjust POT9 on MP3.
d) If the mAs still beyond the 10% range, for 80kV and 110kV, re-adjust POT2 on MP4 a
little, for 40kV re-adjust POT3 on MP4 a little.
3. Power off, take off extended board, connect jumper W.
The mAs meter must be connected tightly, otherwise, it may cause injury or
WARNING damage. This test is only performed in the factory when it is required by
U
Page 4-8
Calibration
4.4.1 Measurement
The X-ray generator is composed of high voltage transformer, high voltage rectifier and X-ray
tube in one pressurized oil tank. The X-ray tube voltage can be measured with indirect method
by the kVp meter.
Load factor Fluoroscopy: 110 kV/4.0 mA, 99 kV/3.4 mA; 66 kV/1.6mA
Radiography: 110 kV/2 mAs,99 kV/2 mAs, 66 kV/2 mAs
4.4.2 Adjustment
1. Pre-adjustment
a) Set Digital Volt-Ohm-Meter meter to DC.
b) Connect 1 probe to TP1 on MP3 PCB, the other to the ground.
c) Power on the system.
d) Set the kV value to 40KV, the reading on the meter should be 4V±0.2V, otherwise
adjust POT2 on MP3 board.
e) Set the kV value to 110KV, the reading on the meter should be 11V±0.2V, otherwise
adjust POT1 on MP3 board.
2. Put kVP meter 50cm or closer (depend on the kV settings) from Focal Spot of the tube.
3. Fluoroscopy kV adjustment
Set the values as below:
KV Display 66kV 99kV 110kV
Measurement
The errors of the KVP should be within ±7% of the values above, Otherwise adjust POT116 on
MP5. (Counter-clockwise to increase kV, and clockwise to decrease kV).
Also monitor DC Voltage on MP6 (+RPHV and –RPHV), the trend is as same as kV trend,
it should be no more than 370V.
4. Radiography KV adjustment
Select the Radiography mode and set 2mAs, press the HandSwitch to verify the test
value according to the table below.
Page 4-9
EverView 7500 C-Arm X-Ray Imaging system
The errors of the KVP should be ±7% of the KV values above, Otherwise adjust POT2 on
MP5. (Counter-clockwise to increase kV, and clockwise to decrease kV)
Also monitor DC Voltage on MP6 (+HRPV and –HRPV), the trend is the same as kV trend,
it should be no more than 395V.
Before perform the Radiography KV adjustment procedure, you must adjust
NOTE
the Fluoroscopy KV. Step 3 and Step 4 must not be reversed.
Page 4-10
Calibration
Page 4-11
EverView 7500 C-Arm X-Ray Imaging system
Page 4-12
Calibration
Brake Cover
Fastening Adjustment
Screw Screw
Page 4-13
EverView 7500 C-Arm X-Ray Imaging system
Limited
Screws
Page 4-14
Calibration
Fixed
Screw
Page 4-15
EverView 7500 C-Arm X-Ray Imaging system
3. Then you can find a hole used to lock the C-Arm rotation, there is a pin extended into
the hole. To repair this equipment, you must use a stick to push the pin back and
install the push screw. See the follow Fig.
Page 4-16
Calibration
2. Take off these two limited screw and then you can loose all fixed screws, turn the nut to
adjust the brake force.
Page 4-17
EverView 7500 C-Arm X-Ray Imaging system
2. Install the alignment tool on the I.I., adjust the center of the tool. Rotate the 2 screws to
adjust the position of the center on the monitor .Select manual Fluoroscopy mode and
take exposures. Check the image on the left monitor and adjust the center of the tool
superposed with the center of the washer.
Page 4-18
Calibration
4. Select Manual Fluoroscopy mode and MAG2 field ,take an exposure. Use an erasable
marker to make a dot on the monitor at the center of the tool. Rotate the camera to 90
°, 180°, and 270°. Mark the tool target center at each camera rotation angle.
5. Draw a line between the 0° mark and the 180°mark. Then draw a line between the
90° mark and the 270° mark. The intersection of the lines forms a cross that
identifies the center axis of rotation.
6. Loosen the three securing screws of the CCD and adjust the center of the CCD to
superpose with the mark. Secure the 3 CCD screws once you have centered the
washer.
7. Take off the washer from the I.I.
Page 4-19
EverView 7500 C-Arm X-Ray Imaging system
4. Adjust the position of the collimator and make the center of the collimator (image on
the monitor) superposed with the center of the washer.
Rotate the 2 screws below to adjust the position of the collimator.
5. Remove the alignment tool. Place a washer on the center of the I.I. And tighten it with a
transparent adhesive tape.
6. Close the shutter to the smallest. Select manual Fluoroscopy mode and take exposures.
The image of the shutter will display on the left monitor. Rotate the shutter to the
vertical position. Reverse the image and check if the center of the shutter superposed
with the center of the washer. Otherwise, adjust the 2 screws of the shutter to move
the center to superpose with the center of the washer. Rotate the shutter to the
horizontal position, and adjust the 2 screws to make the center of the shutter
superposed with the center of the washer.
Page 4-20
Calibration
4.9 Others
For Camera iris adjustment (dose adjustment), Blanking adjustment, -ABS window adjustment,
Refer to Appendix chapter 11 for details.
Page 4-21
Chapter 5
Software
EverView 7500 C-Arm X-Ray Imaging system
5.1.1 Overview
Using EverView 7500 maintainer, the service engineer can:
Page 5-2
Software Maintenance
1. Press ‘K’, ‘Y’, ‘S’ keys orderly and quickly when 7500 software is running in VIEWER screen.
LOGON DIALOG
SELECT USERS
NOTE
The logon password for user SVC is “zhq”, PLEASE KEEP it CONFIDENTIAL.
Page 5-3
EverView 7500 C-Arm X-Ray Imaging system
MAIN WINDOW
Page 5-4
Software Maintenance
• Version of database
• Version of 7500 software
• Version of saved patients
• Amount of saved images
• Amount of EverView 7500 Maintainer
NOTE
Software version takes form of X.YY, in which X means the major version, and
YY means minor version.
To get Version of EverView 7500 Maintainer, click red ‘+’ on the top left corner of the Maintainer
dialog, and then click ‘About 7500SVCTool…’ A message box will pop up to display Version of
EverView 7500 Maintainer.
Page 5-5
EverView 7500 C-Arm X-Ray Imaging system
Page 5-6
Software Maintenance
After press the Repair User Data button, a message box will pop up if the database has crashed.
Press Yes button to repair it.
Page 5-7
EverView 7500 C-Arm X-Ray Imaging system
Page 5-8
Software Maintenance
5.2.1 Overview
There are two CD disks for every 7500 system: Application CD and Recover CD.
NOTE
Only authorized maintaining engineers can use those disks.
Application CD Recover CD
Page 5-9
EverView 7500 C-Arm X-Ray Imaging system
6. The system will ask you if you want to keep user data. The default value is Yes. If you do not
want to save the data, you can select No. Click on Next button, the installation will start.
If you select No, the images and patient information saved in the system will be
NOTE
lost.
7. After the installation is finished, the system will give a Success! prompt. You can click on
Finish button. The software is installed successfully.
Page 5-10
Software Maintenance
8. Wait for a few minutes, the DaoJet database installation Wizard will start automatically.
Page 5-11
EverView 7500 C-Arm X-Ray Imaging system
9. Select the default settings and click Next button to go all the way down the install process
until the screen below pop up. Click on OK button to continue.
10. After the installation is finished, the following screen will pop up.
Page 5-12
Software Maintenance
5.2.3 Recover CD
DESCRIPTION: The recover CD can only be used to repair the operating system when necessary.
Steps:
1. Press Alt + Ctrl + Delete -> Click ‘Shutdown’ -> Choose ‘Restart’;
2. Click ‘Start menu’ -> Click ‘Shutdown’ -> Select ‘Restart’;
3. Press Alt + F4 -> Click ‘Shutdown’ -> Select ‘Restart’;
4. Click ‘Start menu’ -> Click ‘Logoff’ -> After logging off, click ‘Shutdown’ -> Select ‘Restart’.
If any method you are trying can't work, please try another one.
Page 5-13
Chapter 6
Trouble Shooting
EverView 7500 C-Arm X-Ray Imaging system
4. Fault indicator
A). If Fault Indicator is not lit, go to the next step;
B). If Fault Indicator is lit:
a) Press the Emergency Stop Switch, and open the main control cabinet;
There is the risk of electric shock when implementing the following steps.
CAUTION
Take appropriate precautions.
Page 6-2
Trouble shooting
d) The operator needs power off the system and wait for 1minute, then power on the
system to reset the Alarm.
6.workstation check
A). If workstation works normally, go to the next step;
B). Check and verify the monitor is powered on.
CAUTION The following step has the risk of electric shock. Take appropriate precautions.
Page 6-3
EverView 7500 C-Arm X-Ray Imaging system
Begin
Yes
Yes
No
No
Other indicators
correct? Check MP2, MP3 and display PCBs
Yes
No
Display of the Check monitor power cable, filter,
monitors correct? computer power supply
Yes
End
Page 6-4
Trouble shooting
Page 6-5
EverView 7500 C-Arm X-Ray Imaging system
POT 2 POT 1
TP1
Page 6-6
Trouble shooting
Before follow the following step, the maximum DC voltage might exceed 380V,
CAUTION
take appropriate precautions.
Page 6-7
EverView 7500 C-Arm X-Ray Imaging system
System Check
Page 6-8
Trouble shooting
N Check
kV/mA is
Display PCB
OK?
and MP2、MP3
Y
Measure the MP3 KVC
signal
1V=10kV
N Check
KVC: 4--11V?
MP3
Y
Select Radiography
Check
N
Display
kV is ok?
MP2 and
MP3
Y
Page 6-9
EverView 7500 C-Arm X-Ray Imaging system
N Check display
mAs Range is
board
ok??
Check MP2
Y
Power down
Page 6-10
Trouble shooting
Electrical circuits inside the equipment use voltages that are capable of causing
WARNING
serious injury or death from electrical shock. Take appropriate precautions.
Page 6-11
EverView 7500 C-Arm X-Ray Imaging system
b). The value of wave is about 0.6V, or readjustment shall be made according
to the current calibration procedure in Chapter 5. The wave should be as
the following.
Page 6-12
Trouble shooting
Page 6-13
EverView 7500 C-Arm X-Ray Imaging system
Page 6-14
Trouble shooting
Fluoroscopy Check
Begin Fluoroscopy
check
Attach working
card
Power on
use Manul mode
The monitor
brightness is N Check the IBS signal
suitably and the in MP3
kV is stably?
Y
5
End
Page 6-15
EverView 7500 C-Arm X-Ray Imaging system
N Check
It is correct?
MP5,MP7
Y
Use a load instead of the
tube HV primary,Check
the wave of the HV
N Check MP5,
The high voltage
MP6, MP7 and
wave is correct?
Motherboard
Y
Check the
tube
assembly
Page 6-16
Trouble shooting
Check the
image chain
Check
the II
IBS:0--12V,is
fluctuate when the Check the
brightness is cable of II
changing?
Check
MP3
Page 6-17
EverView 7500 C-Arm X-Ray Imaging system
Electrical circuits inside the equipment use voltages that are capable of causing
WARNING serious injury or death from electrical shock. Take appropriate precautions.
1. In normal condition
• Insert the MP3 board to mother board via extended board;
• Use an oscilloscope to measure the voltage of W test point on extended board.
• Power on and select radiography mode.
• Adjust KV value to 70 and mAs to 20.
• Press handswitch to have an exposure, the exposure time is about 1 second and
the test result is approximate –5V.
• Power off the system.
2. In abnormal condition
A). An alarm is given from the fault indicator during an exposure.
a) Shut off the system and check the tube’s temperature. If the temperature is too
high (the temperature of tube cannot exceed 55℃, or an alarm will be given),
re-start the machine 20 minutes later and make the exposure inspection again.
b) MP2 board is connected to motherboard via connection board.
c). Power on and take an exposure, if an alarm is given again, check the signal
“FILAL-HC, AL, ALTH-HC, HVTDV” on extended board.
d). FILAL-HC is filament over-current alarm signal, AL is Hi-voltage primary
over-voltage alarm signal, ALTH-HC is overheating alarm signal of tube, and
HVTDV is Hi-voltage primary low-voltage alarm & sampling signal.
B). The voltage tested on the oscilloscope is far away from 5V;
Readjust it in line with description in Chapter 4.
C). During the exposure, the buzzer beeps continuously but the alarming indicator is off.
a). Shut off the system for 1 minutes. Then restarts the system and conduct
exposure. If the same problem happens again, open the cover of console under
power-up condition, and adjust the voltage of TP9 on MP7 board to 13.8V, but no
more than 14V.
b). Conduct exposure, if the same problem happens again, check if the fluoroscopy
current is correct according to measuring methods described in Chapter 4.
c). Replace the tube.
Page 6-18
Trouble shooting
D). Others
During the exposure, the exposure lamp shines and the buzzer beeps continuously
at the same time. Use a DC-voltage-meter to check the HDSW-EN signal on MP5.
Value of the signal is about 15V or check the signals (HDSW-EN、HDSW-SECURITY
and H6)on mother board. When H6 is on, check the HANDSWITCH signal.
Page 6-19
EverView 7500 C-Arm X-Ray Imaging system
Begin
Take an exposure
No
Exposure logic correct? Yes Check MP4, MP6
Fault indicator on?
and MP7
No
Yes
Check MP7
Yes
End
Page 6-20
Trouble shooting
Fault 2:
Image looks gloomy
Reason:
Contrast/Brightness parameter is improper.
Repair:
Click right button of mouse when software runs in MAIN screen, then select
Contrast/Brightness item to adjust contrast/ brightness parameters.
Fault 3:
Archive function can’t work.
Reason:
Archive disks damaged.
Repair:
Repair or replace those devices including USB disk, CDRW disk.
Page 6-21
EverView 7500 C-Arm X-Ray Imaging system
All above signals effective will activate two LEDs, one is the Fault Indicator on the control panel
and the other is H1 on MP5 board.
Other way, a buzzer on MP7 board will generate continue “Bi-Bi-….” audible signal for the over
current of the HV transformer primary coil. It indicates that there is a problem on inverter.
The operator should power off the system and wait for 1minute, then power on the system to
reset the Alarm.
Page 6-22
Chapter 7
Parts Replacement
EverView 7500 C-Arm X-Ray Imaging system
Electrical circuits inside the equipment use voltages that are capable of Causing
WARNING
serious injury or death from electrical shock. Use appropriate precautions.
Please use specified fuse.
If the system has power applied, power off the machine and unplug the AC
WARNING power cord from the wall outlet.
Page 7-2
Parts Replacement
Page 7-3
EverView 7500 C-Arm X-Ray Imaging system
Page 7-4
Parts Replacement
Page 7-5
EverView 7500 C-Arm X-Ray Imaging system
1 2 3 4 5 6 7 8
Page 7-6
Parts Replacement
11
10 32
16
Page 7-7
EverView 7500 C-Arm X-Ray Imaging system
40 41 42 33 34
38 39
71 39
Page 7-8
Parts Replacement
37 13 43 36 35
Page 7-9
EverView 7500 C-Arm X-Ray Imaging system
35 36 31
70 14
Page 7-10
Parts Replacement
70 14
29
30
Page 7-11
EverView 7500 C-Arm X-Ray Imaging system
17 18 19 21 20
Page 7-12
Parts Replacement
47 26
67
79
Page 7-13
EverView 7500 C-Arm X-Ray Imaging system
15
60
45
53
22
59
Page 7-14
Parts Replacement
61
25
58
47
69
54
55
57
56
12
69
74
Page 7-15
EverView 7500 C-Arm X-Ray Imaging system
46
48
49
51
52
Page 7-16
Parts Replacement
64
63
75
76
83
82
68
Page 7-17
EverView 7500 C-Arm X-Ray Imaging system
80
50
62
Page 7-18
Parts Replacement
NOTE 2415142 kit includes 2 Power Supply fuses, 2 AC24V fuses and 2 fuse sockets.
Power supply
Fuse(6.3 A)
AC 24V Fuse(1 A)
Page 7-19
EverView 7500 C-Arm X-Ray Imaging system
4. Power on the system; press the semi-transparent shutter close for more than 10s.
5. Check and verify the movement of the leaf is smooth.
6. Select fluoroscopy mode to take exposure, verify the shutters rotation is at the
center of the blanking circle, if not, loosen the screws, which is used to fasten the
shutter, re-position the shutters, until the shutters opening is at the center of the
blanking circle, see above Fig for reference.
7. Fasten locating screws.
MP1
Page 7-20
Parts Replacement
8. Select the radioscopy mode, adjust the resistor POT303 on the MP1 board, let the
collimator leafs open to maximum position and the voltage measured on the two
poles of the drive motor less than 3V.
9. Power off.
10. Connect Iris (see below Fig for reference), and install it on its position but don’t screw
down.
11. Take off the bridge gear, close the IRIS manually, then power on the system.
12. Press the IRIS close Icon for more than 10s, and then adjust the resistor on the IRIS
until the motor don’t rotate. Keep the motor resting and install the bridge well.
13. Select the manual fluoroscope mode and take exposure, verify the IRIS field is at the
center of the blanking circle. If not move the position to adjust. Screw down the
fastening nuts.
14. Open the IRIS to maximum, take exposure and adjust the POT302 on the MP1 board
to make the X-ray field in tangent with blanking circle.
15. Place the Beam Alignment Fixture Tool (9 inch) in front of the Image Intensifier, close
the IRIS to minimum, take an exposure to measure the minimum field size is less
than 45mm, if not repeat the step14 and step 16.
16. Select MAG1 field mode, take exposure and adjust the POT2, to make the X-ray field
in tangent with blanking field.
17. Select MAG2 field mode, take exposure and adjust the POT1, to make the X-ray field
in tangent with blanking field.
18. Select the radioscopy mode, adjust the resistor POT304 on the MP1 board let the IRIS
open to maximum position but the voltage measured on the two poles of the drive
motor less than 3V.
19. Power off, take off extended board, fix pressing strip, and recover the console.
Page 7-21
EverView 7500 C-Arm X-Ray Imaging system
Page 7-22
Parts Replacement
Page 7-23
EverView 7500 C-Arm X-Ray Imaging system
4. Connect a DC voltage meter (500V) to +PRHV and -PRHV on MP6 board, power on, the
value is about 140V.
5. Press kV+, the voltage rises smoothly. When 110kV is displayed, adjust POT116 to make
the voltage about 360V.
Page 7-24
Parts Replacement
6. Switch to radiography mode, and then adjust POT2 to make the voltage be 370V.
7. Adjust kV according to chapter 4.4.
8. Power off. Install the strip pressing the PCB and recover the console.
Page 7-25
EverView 7500 C-Arm X-Ray Imaging system
3. Remove X-Ray generator assembly and replace a new one, when you install the
generator, the torque to four fastening screws is 5N*M.
4. Power on, select manual fluoroscope mode.
5. Adjust fluoroscopy current, fluoroscopy kV, radiography mAs, radiography kV according
to chapter 4.
6. Power off the system.
7. Cover the X-ray generator assembly.
8. If the X-Ray generator has been produced for more than 3 months, implement a warm up
procedure according the procedures described in chapter 4.5.
Page 7-26
Parts Replacement
BIOS Setup
1. Press“F2/DEL”key to enter BIOS。
2. Set up “EXIT/Load Optimized Defaults ok”, Enter.
3. “Main/System Time” Set up the time now.
“Main/System Date” Set up date now
4. Advanced/Floppy Configuration/ Diskette Controller [disable]
5. Advanced/Boot Configuration/Num Lock [off]
6. Power/After Power Failure [power on] [automatically power on when the power is
available]
7. Boot /Silent Boot [Disabled] [Disable Tongfang LOGO]
8. Boot / Boot Device Priority/1 st Boot Device [Harddisk]
2 nd Boot Device [CD driver]
3 rd Boot Device [U-Stick]
9. Exit/ Exit &Save Changes Press “y ”to exit.
Page 7-27
EverView 7500 C-Arm X-Ray Imaging system
Page 7-28
Parts Replacement
5) Turn over the console of C-Arm slowly (3 men) to put it on the floor.
Try to lay it down slowly to prevent quick turnover and hurt to person.
WARNING
6) Remove the cover on the Electric Cylinder and save the screws and cover.
Page 7-29
EverView 7500 C-Arm X-Ray Imaging system
Page 7-30
Parts Replacement
7) Remove the locking screw-plug on the pin and draw out the pin as picture below.
Locking washer
Page 7-31
EverView 7500 C-Arm X-Ray Imaging system
9) Screw the M4X30 Screw in the hole on the axis, and then use a clamp to draw the pin out.
10) Unplug the housing of the power supply wires of the Electric Cylinder. And cut off the 2
plastic strapping. Save the housing for new Electric Cylinder wires.
Page 7-32
Parts Replacement
11) From the bottom of the C-Arm console, draw the Electric Cylinder assembly out.
2) Insert the pin with the M4X30 screw into the hole, and check if the pin is in the axis of the
Electric Cylinder assembly.
Page 7-33
EverView 7500 C-Arm X-Ray Imaging system
3) Unscrew the M4X30 screw and tighten the nut on the hole.
4) Put the locking pin on the bottom of the Electric Cylinder and lock the pin with the locking
screw-plug. Put the Electric Cylinder cover back and tighten it with screws.
5) Lift the C-Arm and make it stand on the floor. Reassembly all the removed parts of C-Arm.
6) Connect the workstation and the footswitch. Power on the system. Check if the C-Arm can
rise and drop freely.
Page 7-34
Parts Replacement
3. Rotate the brake handle to vertical position and remove the 2 screws of the fixing block.
4. Push the rotation brake handle backward with your left hand, remove the block with your
right hand.
Page 7-35
EverView 7500 C-Arm X-Ray Imaging system
5. Rotate the brake handle to make the fixing screw upward, unscrew the fixing screw and
remove the handle.
6. Grasp the left end of the brake shaft, knock on the right end of the shaft, remove the bearing
bushing on the left, then remove the shaft from the shaft from the right side. Refer to the
pictures below.
7. Get the new brake assembly, remove the handles, and put the shaft in position as the
pictures below.
Page 7-36
Parts Replacement
9. Rotate the handles to the upward position ( as pictures below), put the block in position and
fasten the screws of the block.
NOTE
While putting the block, shake the shaft with your left hand to make it in position.
10. Rotate the brake handle to lock position, then release the brake. Verify if the brake can be
operated smoothly. Release the brake, rotate the C-Arm, then lock the brake. And check if the
brake force is enough. If the brake force is not enough, follow the step 11 to adjust the brake
force; Otherwise please jump to step 12.
Page 7-37
EverView 7500 C-Arm X-Ray Imaging system
11. If the brake force is not enough , or it is too large, release the brake. Remove 3 of the 4 screws
as pictures below and keep the one on the top left corner. Rotate the brake flake with the
screw. Rotate clockwise to decrease the brake force; Rotate counter-clockwise to increase
the brake force.
12. After the brake force is appropriate, fasten the 4 screws of the brake flake. Check and verify
the operation of the brake. Reassemble the cover of the brake, fasten the screws of the
cover.
Page 7-38
Parts Replacement
2. Unscrew the 2 fastening screws of the brake plate, then you can lift the brake assembly.
3. Then you can replace the old assembly with a new one. After that, check and verify that if the
brake force of the orbital brake is appropriate. If not, please adjust the brake force as step 4.
Otherwise you can jump to step 5.
Page 7-39
EverView 7500 C-Arm X-Ray Imaging system
4. Loosen the 2 fastening screws, fasten or loosen the 4 adjustment screws to adjust the brake
force. Rotate the screws clockwise to decrease the brake force, rotate them
counter-clockwise to increase the brake force. Check and verify that the brake force is
appropriate.
Fastening
Screw(2)
Adjustment
Screws( 4)
Page 7-40
Parts Replacement
2. Remove the limited screws on the shaft. Loosen all of the fixing screws of the brake nut.
Remove the brake handles.
Fixing Limited
screws screws
Page 7-41
EverView 7500 C-Arm X-Ray Imaging system
4. Adjust the nut on the shaft to get the appropriate brake force. Rotate it clockwise to increase
the brake force, rotate it counter-clockwise to decrease the brake force.
Nut
5. Reassemble the brake assembly and cover and fasten the screws.
Page 7-42
Parts Replacement
2. You can adjust the brake force by adjusting the 4 screws of the brake plate.
3. You can remove the old brake plate and replace it with a new one.
4. If the brake handle is not vertical when in lock state, you can loosen the nut below to adjust
the handle to vertical position.
Page 7-43
EverView 7500 C-Arm X-Ray Imaging system
Page 7-44
Parts Replacement
1 GND GND GND GND GND GND GND GND GND GND HVTN HVTN HVTGND HVTGND A
2 15V 15V 15V 15V 15V 15V 15V 15V 15V 15V HVTN HVTN HVTGND HVTGND B
3 -15V -15V -15V -15V -15V -15V -15V -15V -15V -15V C
6 IRD IRI ALTH-LC HVCOM-HC MAC KVC MAC KVC KVC HVCOM-HC HVTP1 HVTP1 F
7 OPEN CLOSE MNSYNC GENRES-HC IBS GENRES-HC GENRES-HC GENRESHC MNSYNC HVTP HVTP H
15 SWMT1 SWMT2 MAS2 MAS1 DATA6 DATA5 FILLA1 FILLA1 GND GND S
17 SRMT1 SRMT2 MAS6 MAS5 FILSM2 FILSM2 FFOND FFOD PWRSMDV HVTDV CL13N CL13 U
18 FTALR MAS7 FILC0M FILC0M GND GND DISCOM-HC RECTTR FFOD FFO V
HEATALM
19 FSMT1 FSMT2 HDSW-LC FTSW-LC GND MAMEAS2 CL13N CL13 _HC FFOND FFON W
21 KEY KEY Emg_swt 0VREF KEY KEY KEY KEY KEY KEY KEY KEY KEY KEY Y
22 GND GND GND GND GND GND GND GND GND GND GND GND GND GND Z
Page 7-45
Chapter 8
Periodic Maintenance
EverView 7500 C-Arm X-Ray Imaging system
Page 8-2
Periodic Maintenance
After taking X-rays, the high voltage cables can retain a lethal charge, even if
WARNING you power down the system. Even though the contacts are recessed, high
voltages can arc several inches from their terminals if the cables are removed
from the connector wells. Use caution when disconnecting high voltage
cables even when the power is turned off. If you remove these connections,
immediately discharge their contacts against the high voltage tank case
(ground).
Many of the PCB's in this system contain components which are sensitive to
CAUTION
Electro-Static Discharge (ESD). Observe ESD safety procedures.
Page 8-3
EverView 7500 C-Arm X-Ray Imaging system
8.5.1 Tools
Common hand tools
Page 8-4
Periodic Maintenance
8.6 PM Checklist
Serial No: GE HuaLun Medical Systems, Inc
Start Date: EverView 7500 C-Arm X-Ray Imaging
System
GE HuaLun Elapsed Time Meter:
Periodic Maintenance Inspection
PM Schedule: Annual
Report
SEMI-ANNUAL PM
1.1Mechanical Inspections 1.2Electrical Inspections
□Brakes □Horizontal Movement □Electrical Plug/Power Cord
1.SAFETY
GND Pin
□Radial C-arm Rotation
□Ground Continuity: Ohms(highest only)
□Orbital C-arm Rotation
2.1 Line Regulation 2.2 Switch-mode Power Supply
□No-Load Line Voltage(VN): VAC Check and Verify the outputs of the Power
Supply:
□Loaded Line Voltage:
DC +15V
80mAs@100kVp=(VL) VAC
DC -15V
2.POWER SUPPLIES
Page 8-5
EverView 7500 C-Arm X-Ray Imaging system
ANNUAL PM
4.1Radiographic Accuracy (±7%)
□kVp Accuracy @ 2mAs : 60kVp ,90kVp ,110kVp
4.2Fluoroscopic Accuracy (±7%)
□kVp Accuracy: 60kVp ,90kVp ,110kVp
4.ACCURACY
Page 8-6
Periodic Maintenance
COMMENTS
Page 8-7
EverView 7500 C-Arm X-Ray Imaging system
Page 8-8
Periodic Maintenance
Electrical circuits inside the equipment use voltages that are capable of
WARNING
causing serious injury or death from electrical shock. Use appropriate
precautions.
1. If the system has power applied, place the C-Arm power switch in the OFF position and
unplug the C-Arm's AC power cord from the wall outlet.
2. Remove C-Arm and Workstation covers as needed.
• Ground Continuity
Do not perform this continuity test with the AC plug plugged into a power
WARNING
receptacle.
Page 8-9
EverView 7500 C-Arm X-Ray Imaging system
From To Spec
Pin 4 of the FootSwitch Connector Screw on the bottom of the < 0.5 Ω
FootSwitch
2. Connect the Integrated Cable to the C-Arm and verify continuity between the
following:
From To Spec
Screw on the top left corner of the WorkStation AC Power Plug Ground Pin < 0.5 Ω
Screw on the tube cover AC Power Plug Ground Pin < 0.5 Ω
Screw on the bottom of the FootSwitch AC Power Plug Ground Pin < 0.5 Ω
3. Record only the highest resistance on the PM form.
8.8.4 PC Clean
1. Remove the workstation cover.
2. Take the PC out of the workstation.
3. Remove the PC cover.
4. Clean the PC outside and inside.
5. Recover the PC cover and put it back into the position, recover the workstation cover.
8.9.1 Power ON
1. Connect the integrated cable in the C-Arm connector and lock it. Connect the
FootSwitch cable to the C-Arm.
2. Plug the AC power cord into the wall receptacle.
3. Press the system on key located on the C-Arm control panel.
Page 8-10
Periodic Maintenance
1. Take X-Ray exposure and verify the following Image Orientation functions:
A. Image rotation key rotates the image clockwise and counter-clockwise.
B. Image Invert key inverts and reinverts the image.
2. Press the field size key to change the field size: NORM, MAG1 and MAG2.
3. Using the Converging Line pairs/millimeter (lp/mm) tool on the Image Intensifier,
verify the resolution in lp/mm is equal to, or greater than, the values listed below:
Indicate your measurements on the Checklist.
4.5″ II 6″ II 9″ II
MAG2 ≥ 2.2 MAG1 ≥ 1.8 NORM ≥ 1.2
4. Press the F3 key to verify the image on the left monitor shifts to the right monitor. Verify
the resolution in lp/mm is equal to, or greater than, the values listed above.
5. Press each of the following collimator keys and verify the correct collimator function:
A. Press the COLLIMATOR IRIS OPEN/CLOSE key.Press the Footswitch or HandSwitch and
verify that the size of the circle varies as the iris is opened and closed.
B. Press the COLLIMATOR LEAF OPEN/CLOSE key and press the Footswitch or
HandSwitch. Verify that the distance between the two lines increases and decreases
as the leaves are opened and closed.
C. Press the COLLIMATOR LEAF ROTATE key.Press the Footswitch or HandSwitch and
verify that the image displays on the Workstation monitor and that the icon rotates
as the collimator leaves are rotated.
6. Press the kVp key and verify that the technique can be adjusted manually.
7. Select Auto Fluoro mode and the NORM field size. Verify that the kVp tracks accurately as
copper filters are inserted into the field per the following ranges (9-inch). Indicate your
measurements on the Checklist.
8. Select High Level Fluoro (HLF) mode via the HLF key on the C-Arm control panel.
A. Cover the II with the lead blocker and press the Foot/handswitch to take 1s exposure.
Page 8-11
EverView 7500 C-Arm X-Ray Imaging system
Page 8-12
Periodic Maintenance
WARNING The following steps produce X-rays. Use appropriate safety precautions.
Page 8-13
EverView 7500 C-Arm X-Ray Imaging system
Page 8-14
Periodic Maintenance
1. With the C-Arm's wheel brake engaged, turn the right steering handle 90° to the left and
right. Verify that the wheels move 90° to the left and right.
2. Release the C-Arm's wheel brake. Push the C-Arm slowly and verify that when the steering
handle is turned the C-Arm turns.
Page 8-15
EverView 7500 C-Arm X-Ray Imaging system
3. Lock the wheel brakes completely. Push the Workstation and verify that the casters lock in
place and the wheels roll in a straight line.
4. Release the Workstation's wheel brake. Push the Workstation and verify that the wheels
and casters turn freely.
8.13.2 Power ON
Complete the Power On test.
1. Apply Power to the system and verify that the Workstation and C-Arm complete the boot
sequence and are functional.
2. While pressing the Emergency Stop switch, verify that vertical column operation and X-ray
generation are disabled.
4. Reboot the system by inserting the share key in the hole of the Emergency Stop Switch,
and rotating it (Rotate the emergency pushbutton upward) in the direction according to
the arrow on the pushbutton. Shut down the workstation, then press the Power Off key.
Wait for a few minutes and then back to Power On key on the C-Arm.
5. Verify that the system boots-up error free.
Except for the most urgent situation, workstation should be shut down first
NOTE
(Pressing F12 on the keyboard) before cutting off the power even if the
emergency button is pressed down.
Page 8-16
Chapter 9
Circuit Schematics
EverView 7500 C-Arm X-Ray Imaging system
List
Page 9-2
Circuit Schematics
Page 9-3
EverView 7500 C-Arm X-Ray Imaging system
Page 9-4
Circuit Schematics
Page 9-5
EverView 7500 C-Arm X-Ray Imaging system
Page 9-6
Circuit Schematics
Page 9-7
EverView 7500 C-Arm X-Ray Imaging system
Page 9-8
Circuit Schematics
Page 9-9
EverView 7500 C-Arm X-Ray Imaging system
Page 9-10
Circuit Schematics
Page 9-11
EverView 7500 C-Arm X-Ray Imaging system
Page 9-12
Circuit Schematics
Page 9-13
EverView 7500 C-Arm X-Ray Imaging system
Page 9-14
Circuit Schematics
Page 9-15
EverView 7500 C-Arm X-Ray Imaging system
Page 9-16
Circuit Schematics
Page 9-17
EverView 7500 C-Arm X-Ray Imaging system
Page 9-18
Circuit Schematics
Page 9-19
EverView 7500 C-Arm X-Ray Imaging system
Page 9-20
Circuit Schematics
Page 9-21
EverView 7500 C-Arm X-Ray Imaging system
Page 9-22
Circuit Schematics
Page 9-23
EverView 7500 C-Arm X-Ray Imaging system
Page 9-24
Circuit Schematics
Page 9-25
EverView 7500 C-Arm X-Ray Imaging system
Page 9-26
Chapter 10
Technical Reference
EverView 7500 C-Arm X-Ray Imaging system
Page 10-2
Technical Reference
10.13 Tolerance:
The tolerance of the tube voltage ±7 %
The tolerance of the tube current ±10 %
The tolerance of the current time product ±10 %
Page 10-4
Technical Reference
Patients must be kept as far as possible from the focus, with minimum
WARNING distance of 20cm. Make sure not to operate when a plastic cover in x-ray
assembly is disassembled, which properly make the distance meet the above
requirements.
The operating table matched this equipment should be made of small
NOTE
attenuation equivalent material, otherwise, X-ray dose may be increased.
EverView 7500 C-Arm X-Ray Imaging system
Page 10-6
Chapter 11
Appendix
EverView 7500 C-Arm X-Ray Imaging system
Counterweight
Filter Grid
11.1.1 Components
The X-Ray Imaging Unit consists of the following items :
1 2
5
3
4
Page 11-2
Appendix
AIR FLOW
CCU
Page 11-3
EverView 7500 C-Arm X-Ray Imaging system
The cable between the camera head and the CCU must be shielded to be in compliance with
the electromagnetic standard IEC 60601-1-2
The shielding of the cable must be connected to the housing of the Image Head and to the
housing of the CCU.
Page 11-4
Appendix
1
9
2
10
3
11
4
12
5
13
6 control input 3 (UC3)
control input 2 (UC2) 14
7 control input 1
15
8 + 24 V DC power in (UE)
power 0 volt (0VP)
(1) Open the circular top cover of the Image head protecting the HVPS (2 screws).
(2) Connect the High-Voltage Power Supply interface as follows (a connection kit is
provided).
UE (15) 24 V DC
HV3-3 0VP (8)
N
UC1 (14) M1
UC2 (6)
UC3(13) M2
Page 11-5
EverView 7500 C-Arm X-Ray Imaging system
Page 11-6
Appendix
1
+ 24 V DC 14
2 0 V - ground
AGC off 15
3 ABC -
ABC + 16
4 Analog ABC voltage
Image rotation 17
5 Image rotation
Last Image Hold 18
6 DRF on / off
Recursive filter 0 19
7 Recursive filter 4
Recursive filter 8 20
8 Recursive filter 16
Horizontal inversion 21
9 Vertical inversion
Freeze A 22
10 Freeze B
23
11
24
12
25
13
Note : some of these contacts are unused, depending upon camera configuration.
Page 11-7
EverView 7500 C-Arm X-Ray Imaging system
4. Protective earth
Three connection points for Protective Earth are available on the RIU:
• One is located near the camera head,
• An other one is located on the HVPS,
• An other one is located on the CCU.
Page 11-8
Appendix
LIH (5)
Analog
ABC (16) DRF (18)
V inv (21)
ABC- (15)
RF 0 (6) Video 3
(100 or 120 Hz)
RF 4 (19)
RF 8 (7)
ABC
RF 16 (20) 10-bit D/A
converter
Image Image
processing : 50 (60) buffer Freeze A (9)
CCD Video 10-bit A/D to
sensor amplifier converter DRF 100 (120)
LIH Hz
Image
H&V inv converter Freeze B (22)
buffer
6. Video 2 output
(1) The Video 2 output displays the image present on the CCU bus. It could be
either the live image coming from the CCD (Last image Hold =1) or a stored
image coming from the image memory (Last image Hold =0).
(2) The live image displayed through the Video 2 output includes noise reduction
feature with the motion detection.
7. Video 3 output
(1) While Freeze A = 1, the Video 3 output displays the image present on the CCU
bus. This image can be horizontally or/and vertically inverted using the H&V
image inversion control inputs.
(2) The image is frozen on the Video 3 output when Freeze A = 0.
8. Video 4 output
(1) The Video 4 output works like the Video 3 output with the Freeze B control input
instead of Freeze A.
(2) The H&V image inversion control inputs act simultaneously on both channels A
and B. They are not active on frozen images.
Page 11-9
EverView 7500 C-Arm X-Ray Imaging system
Interface to
the equipment
The Freeze A and the Last Image Hold inputs used simultaneously allow to freeze the live
image (LIH).
Remark: replace Freeze A by Freeze B in case of the monitor is connected on the Video 4
output.
Page 11-10
Appendix
Interface to the
equipment Monitor A : Monitor B :
live image stored images
+
Last Image Hold
The Freeze A and the Last Image Hold inputs used simultaneously allow to freeze the live
image (LIH).
The Freeze B input is used to maintain the stored image on the monitor B after reading.
The Last Image Hold input is used to maintain the last image on the monitor A when used in
combination with the Freeze A input or to read a stored image.
Page 11-11
EverView 7500 C-Arm X-Ray Imaging system
11.1.4 ADJUSTMENTS
1. System adjustment location
The RIU includes 3 kinds of adjustments:
- The mechanical adjustment of the camera head (centering, focus, iris) ;
- The electrical adjustment of the Camera Control Unit (video gain, white level, video
mask ...) ;
- The electrical adjustment of the HVPS (U1, U2 and U3 voltages).
camera HVPS
head adjustment
adjustment
Camera
Control Unit
adjustment
Page 11-12
Appendix
Without its top cover, the radiological imaging unit is not in compliance
with x-ray DHHS regulation during the adjustments of the camera.
Page 11-13
EverView 7500 C-Arm X-Ray Imaging system
output image
of the XRII
CCD
active area
NOTE
The required torque is 4 N.m in case of this camera with image rotation (M5
screws).
Mechanical focusing
Normally, the factory adjustment should not be modified, except in case of camera
replacement. The focusing procedure is as follows:
1- Open the camera head top cover (remove the 2 M2.5 screws for cameras with
image rotation).
2 x M 2.5 screws to
3 x M5 screws for remove the camera
camera head head cover
centering
Page 11-14
Appendix
The camera must imperatively be turned off before removing the top cover of
NOTE
the camera head (risk of short circuit).
Two mechanical adjustments are available on the top of the camera head as shown on the
following picture:
Mechanical focusing
Iris adjustment
adjustment
close open
focus
VCCD
Test point 1
Page 11-15
EverView 7500 C-Arm X-Ray Imaging system
Iris adjustment
As it is done in the factory, for an input dose of 30 µR/s (75 kVp - 2 mm Cu - uniform X-ray flow)
in the entrance plane of the Radiological Imaging Unit, in normal mode, the iris has to be
adjusted to obtain a CCD signal of 1.3V on the test point TP1.
Use a X-Ray dosemeter to know the X-Ray dose.
In addition, the Automatic Brightness Control (ABC) was adjusted, to maintain the average level
in the center of the CCD at 1.3 volts measured on test point TP1 - see 6.3.4
1- See the previous picture for the iris adjustment location.
2- Use a small screwdriver to adjust the iris aperture.
To recover the factory iris adjustment, see "iris adjustment" on the Thales test
NOTE report. This value indicated the position of the factory adjustment in quarter of
turn with respect to the totally open position.
3. Camera Control Unit adjustment
Page 11-16
Appendix
1
R94 J5
2
P1 P2 P3
J2
J6
1 2 3 4
1
R157
SW2 P4
SW4
2
J4 P11
R156 R163
1
SW3
P5 P6 P8 P9 P10 P7
2 N23 ANB1
The videomask has to be adjusted internally to the image itself ; no edge must be visible inside.
Page 11-17
EverView 7500 C-Arm X-Ray Imaging system
Page 11-18
Appendix
WHEN USING THIS FUNCTION, SOME PRECAUTIONS HAVE TO BE TAKEN TO LIMIT THE
MAXIMUM DOSE IN ORDER TO PREVENT PATIENT INJURY.
Measuring window
U
The measuring window is used for the Automatic Gain Control and for the Automatic
Brightness Control. The average signal in the window is taken into account to control the
video gain in AGC mode and to control the dose (ABC).
The measuring window can be displayed on the monitor using the jumper J3.
The diameter of the measuring window can be adjusted using P5.
The measuring window is centered in the circular blanking. The H & V shift
NOTE
of the circular blanking also move the measuring window.
Analog ABC output
U
The signal in the measuring window (Vlev) is compared to a reference (Vset) in order to
generate the ABC signals.
The analog ABC voltage is available on the pin 16 of J0 interface connector. It varies between
0 V and + 12 V. If the video level in the measuring window (Vlev) is lower than the preset
voltage reference (Vset), the analog ABC voltage is close to 0 V. It means that the entrance
dose is too low: the X-ray generator will increase the dose and the analog ABC voltage will
goes up. If the video level in the measuring window (Vlev) is higher than the preset voltage
reference (Vset), the analog ABC voltage is close to + 12 V. It means that the entrance dose is
too high: the X-ray generator will decrease the dose and the analog ABC voltage will goes
down. The entrance dose will be stabilized at the level which maintain the analog ABC voltage
at about + 6 V.
Page 11-19
EverView 7500 C-Arm X-Ray Imaging system
set-point adjustment
U
Vset is adjusted using P4, ABC set-point. It has been adjusted in our factory to get + 6 V on the
analog ABC voltage output when the video level after the CCD is at 1.3V.
The following figure shows analog ABC voltage versus the entrance dose and the effect of the
6V
Entrance dose
0V
The black level of the video signal in AGC mode is adjusted using R157, black level, on the
N23 ANB1 board.
The amplitude of the video signal in AGC mode is adjusted using R163, white auto level, on the
N23 ANB1 board.
Page 11-20
Appendix
The selection of the J1 output signal is done using the jumper SW4 on the N23 ANB1 board.
4. High-Voltage Power Supply adjustment
HV3-3 HVPS allows the adjustment of U1, U2 and U3 voltages in their respective ranges, to
obtain the 3 different input fields on the XRII.
The HVPS voltages have been adjusted in the factory and, except in special cases, they have
not to be adjusted again.
This adjustment can be done using the adjustment slots on the side of the power supply ; to
have a better access to the potentiometers, it's necessary to dismount the HVPS without
disconnecting the cables.
Normal mode :
U U
Page 11-21
EverView 7500 C-Arm X-Ray Imaging system
NOTE
it's not necessary to adjust U1.
Mag 1 mode :
U U
Page 11-22
Appendix
Page 11-23
EverView 7500 C-Arm X-Ray Imaging system
Page 11-24
Appendix
All the FRUs are delivered with preadjusted parameters : after change, they will be able to
provide an image which will only have to be optimized according to the field constraints.
The parameters which will be optimized will be :
- HVPS replacement : G3 voltages for fields of view, G2 voltage for resolution
- CCU replacement : video mask size and position, possibly shading correction
- camera head : focus, blemishes, centering, iris ; add CCU adjustment (see above)
- XRII replacement : all the above adjustment
G2 (black)
G1 (green)
G3 (red)
G4 (black)
PC (red)
-gauge (blue)
G5 (30 kV)
GND (green-yellow)
Page 11-25
EverView 7500 C-Arm X-Ray Imaging system
Page 11-26
Appendix
Page 11-27
EverView 7500 C-Arm X-Ray Imaging system
G5 (30kV) : push the cable completely in the connection hole, screw the cap carefully.
G1 to G4 cables : push the cable completely in the connection hole, screw the cap.
PC (RED)
-GAUGE (BLUE)
G3 (RED)
GND (GREEN-YELLOW)
G2 (BLACK)
G1 (GREEN)
INPUTS (24V, MODES)
G5 (30KV)
Page 11-28
Appendix
Page 11-29
EverView 7500 C-Arm X-Ray Imaging system
TH 7195 adjustment
After the HVPS replacement, when switching on the RIU, the pre-adjustment of the TH 7195
HVPS for 28HP2 XRII or RIU allows to obtain directly a useful image. But, since every tube and
HVPS are a bit different, it's necessary to adjust gently :
- U2 voltage to optimize the image resolution.
- U3 voltage to optimize the field of view
Use the 3 encoders at the top of the TH 7195
U1 U2 U3
Normal mode :
U U
Mag 1 mode :
U U
Mag 2 mode :
U U
Page 11-30
Appendix
Adjustments :
- video mask / circular blanking : adjust its size and position according to section Video Mask/
U U
anything ; if the edge is globally too dark, a shading correction could be done according to
section Shading Correction Adjustment in 11.1.4, part 3.
use the horizontal and vertical dome corrections gently but check the result for all the image
modes (normal mode, mag1, mag2) ; avoid to use the tilt corrections due to possible rotation
effects.
3. Camera head replacement
Disconnect the camera head cable
Dismount the camera head using the 3 screws on its basis :
Lead sheet
3 screws
Check orientation
between cable,
camera head hole
and HVPS
Dismount the camera head from the orange basis (4 screws) and mount the new one.
Page 11-31
EverView 7500 C-Arm X-Ray Imaging system
4 screws
! The camera head FRU should include a lead sheet on its top
Adjustments :
- Check the image blemishes ; if the image quality is not good enough, dismount the camera
head, clean and mount again.
- centering : using the 3 basis screws, adjust horizontally and vertically the camera head
position according to Camera Head Centering in 11.1.4, part 2.
- iris aperture : it was adjusted in factory using a typical XRII ; if iris has to be adjusted, refer to
IRIS Adjustment in 11.1.4, part 2.
Page 11-32
Appendix
- video mask / circular blanking : adjust its size and position according to Video Mask/ Circular
U U
- Shading adjustment : after image vizualisation, if the uniformity is good enough, don't adjust
U U
anything ; if the edge is globally too dark, a shading correction could be done according to
Shading Correction Adjustment in 11.1.4, part 3.
use the horizontal and vertical dome corrections gently but check the result for all the image
modes (normal mode, mag1, mag2) ; avoid to use the tilt corrections due to possible rotation
effects.
4. XRII replacement
- Disconnect and dismount HVPS and its brackets according to section 11.1.6, part 1 HVPS
Replacement.
- Mount the same camera head according to section 11.1.6, part 1 HVPS Replacement.
- Mount the same HVPS according to section 11.1.6, part 1 HVPS Replacement.
Adjustments :
- adjust camera head and CCU according to section 11.1.6, part 1 HVPS Replacement.
Page 11-33
EverView 7500 C-Arm X-Ray Imaging system
Please read the following instructions as they are designed to prevent damage to property and
harm to the user.
Failure to follow the notes below could result in bodily harm or damage to
WARNING equipment.
• Use only a properly grounded power plug and receptacle. An improper ground may
cause electric shock or equipment damage.
• If your monitor does not operate normally - in particular, if there are any unusual
sounds or smells coming from it - unplug it immediately and contact an authorized
dealer or service.
• Never insert anything into the monitor.
• Do not replace fuse.
Failure to follow the notes below could result in bodily harm or damage to
CAUTION equipment.
• Do not cover the vents on the monitor cabinet. Bad ventilation may cause a breakdown
or fire.
• Put your monitor in a location with low humidity and minimum of dust.
• Place the monitor on a flat and stable surface. The monitor can cause injury by falling.
• When cleaning the monitor case or the surface of the CDT, wipe with a slightly
moistened, soft fabric. Do not clean monitor with water. Use a water-diluted mild
detergent.
• Do not spray detergent directly on the monitor.
• Do not place any heavy objects on the monitor.
• For each hour of looking at the monitor, you should let your eyes rest for 5 minutes.
• Do not use or store inflammable substances near the monitor.
• Keep the monitor away from any magnetic substances.
• Please transmit signals according to the broadcast standard shown in the specification.
Page 11-34
Appendix
11.2.1 Introduction
• This monitor is a very high brightness, high-resolution digitally controlled autoscan
monochrome monitor.
• This monitor, specifically designed for medical applications, gives unique performances
in term of brightness, resolution and spot size.
• A rich set of functions allows quick installation, flexible user preference customization
and easy maintenance.
• Almost all the internal functions of the monitor are digitally controlled, and they can be
stored in a resident memory that may contain up to 18 different formats, 2 of which are
factory preset.
• The monitor complies with the Energy Star Computers Program initiated by the EPA.
• The monitor will automatically synchronize within a wide range of scanning frequencies
and determine the specific operating mode.
• Monitor behaviors, image quality and screen geometry adjustments can be changed or
selected via the hardware OSD by means of the front panel push buttons.
Page 11-35
EverView 7500 C-Arm X-Ray Imaging system
(5) If signal format of system is SOG (sync on green), Signal Selection (switch① in figure 1)
switches to Video.
(6) If signal format of computer is Separate Sync, Signal Selection (switch① in figure 1)
switches to VGA.
4. Setup Daisy Chain Selection
(7) If you use the sole monitor, Daisy Chain Selection (switch④ in figure 1) switches to 75
Ω.
(8) If you use some monitors in Daisy Chain configuration, the units shall be connected as
shown in figure 2. Daisy Chain Selection of the end monitor switches to 75Ω, and the
other monitors switch to High.
Figure1
Connectors on the back of the monitor
Ω Ω Ω
Figure 2
Daisy Chain Configuration
Page 11-36
Appendix
Power On/Off
• Press “ ” to activate OSD menu, then continued press “ ” button more than 2s to
lock keyboard.
Page 11-37
EverView 7500 C-Arm X-Ray Imaging system
Page 11-38
Appendix
Page 11-39
EverView 7500 C-Arm X-Ray Imaging system
Page 11-40
Appendix
• Technical data
Nominal high voltage 110 kV
Nominal anobe input power 2.2 kW
Nominal input power (10 minutes ON) 0.6 kW
Maximum anode heat content 53 kJ
Target material Tungsten
Target angle 10°
Small focal spot 0.6 x 1.4 mm
Large focal spot 1.4 mm
Filtration equiv. 0.55 mm Al
Anode Stationary
Typical filament specification 4.2 V, 3.5 A
Maximum filament current 4.6 A
Weight 0.65 kg
Page 11-41
EverView 7500 C-Arm X-Ray Imaging system
Page 11-42
Appendix
Page 11-43
EverView 7500 C-Arm X-Ray Imaging system
Page 11-44
Appendix
Page 11-45
EverView 7500 C-Arm X-Ray Imaging system
Page 11-46
Appendix
The X-ray generator assembly consists of high voltage transformer, high voltage rectifier
and X-ray tube in one pressurized tank. It can be used in EverView 7500 C-arm X-ray
imaging system.
Page 11-47
EverView 7500 C-Arm X-Ray Imaging system
1. Technical specification
Power:
Standard Fluroscopy(small focal spot)
110kV×4.0mA=440W (maximum)
HLF(small focal spot, pulse)
110kV×8.0mA=880W (maximum)
Radiography(Large focal spot)
110kV×20mA=2.2KW (maximum)
Page 11-48
Appendix
2.Configuration
Time (min)
Reference Axis : The axis from focal spot to the center point of the Image Intensifier.
Note:The focal spot marked in the diagram is the midpoint between large focal spot and
small focal spot, which the windage is no more than 2mm from the benchmark
axes.
3.X-ray tube
Refer to section 11.3.
Page 11-49
EverView 7500 C-Arm X-Ray Imaging system
Fig 1
X-ray tube
Fig 2
Page 11-50
Appendix
b) Connect 9-core connector to glower cable and connect 4-core connector to high voltage
cable.
9-core connector
4core connector
D-sub9 male
D-sub9 female
c) Connect D-sub9 male to Iris and connect D-sub9 female to parallel shutter, please see
picture below for details.
Iris
D-sub9 male
Parallel shutter
D-sub9 female
Page 11-51
Address
SERVICE OFFICES
For service issues, contact the Service Office for your country.
AMERICAS
WAUKESHA, WI 53188
MILWAUKEE, WI 53201
LATIN AMERICA
GE Medical Systems Tel: +1–305–497–1200
USA
ASIA
NIPPON (Japan)
GEMSA Headquarters Tel: 81–426–48–2940
REST OF ASIA
GE Pacific Pte. Ltd (GEMS Asia) Tel: +65–291–8528
SINGAPORE 0316
EverView 7500 C-Arm X-Ray Imaging system
EUROPE
BELGIE/BELGIQUE (Belgium)
GE Medical Systems Benelux NV/SA (NL) Tel:+32 2 644 09 38
DANMARK (Denmark)
GE Medical Systems Danmark Tel: +45 45 51 00 55
DK–2840 HOLTE
DEUTSCHLAND (Germany)
GE Medical Systems Deutschland GmbH & Tel: 49–6102–36–0
Martin–Behaim–Strasse 10
Beethovenstrasse 239
42655 Solingen
E–28100 ALCOBENDAS
FRANCE (France)
GE Medical Systems Tel: +33 (0)1 30 70 40 40
BP34
HELLAS (Greece)
GE Medical Systems Hellas Tel: +30 1 93 24 582
ITALIA (Italy)
GE Medical Systems Italia Tel: +39 02 64 22 01
20126 MILANO
NEDERLAND (Netherlands)
GE Medical Systems Nederland B.V. Tel: +31 73 6 457 457
NL–5231 DD S HERTOGENBOSCH
RײRSTERREICH (Austria)
GE GesmbH Medical Systems Austria Tel: 0660 8651 (geb hrenfrei)
POLSKA (Poland)
GE Medical Systems Polska
PORTUGAL (Portugal)
GE Medical Systems Portuguesa S.A.
ROSSIYA (Russia)
GE Medical Systems Tel:+7 095 935 72 41
Kosmodamianskaya nab. 52, Bldg 1, 6th Floor Fax:+7 095 935 73 46 and 48
Tlx:613020 GEMED SU
SVERIGE (Sweden)
GE Medical Systems Tel: +46 8 457 95 20
S–113 85 STOCKHOLM
SCHWEIZ/SUISSE (Switzerland)
EverView 7500 C-Arm X-Ray Imaging system
CH–6010 KRIENS
T RKIYE (Turkey)
GE Medical Systems Turkiye A.S. Tel: +90 212 75 5552
Gayretteppe
ISTANBUL
UNITED KINGDOM
IGE Medical Systems
SLOUGH
Imagination at work