Conductimetric Analysis of Water Soluble Ionic Contamination of Blasting Abrasives
Conductimetric Analysis of Water Soluble Ionic Contamination of Blasting Abrasives
Conductimetric Analysis of Water Soluble Ionic Contamination of Blasting Abrasives
1
D 4940
ISO method uses strict SI units, this test method reports using SI 8. Calibration and Standardization
compliant units.
Method to Method Comparison:
8.1 Determination of Cell Constant:
The reader is warned that it is difficult to make direct comparisons 8.1.1 The conductivity cell will come with a predetermined
between the results of these two different methods of analysis. constant. This constant should be checked periodically, one
Weight/Volume versus Volume/Volume Method Considerations: method being as follows:
In the ASTM Method the weight of the abrasive is not known; this 8.1.1.1 Prepare a standard solution such as a 0.0005 N
makes it impossible to assess the ratio between conductivity values solution of KCl by diluting a 0.02 N KCl solution with water
determined using this test method procedure and those determined using
or by dissolving 0.0372 g of KCl (heated before weighing for
the ISO 11127-6 procedure.
Comparisons Between Reported Units for Each Method: 1 h at 105°C) in water, followed by dilution to 1 L. Cool and
An independent study by SSPC showed that the relative order of measure the conductance at 25°C as described in Section 9.
extracted salts using each type of procedure on abrasive materials was Calculate the cell constant, K25, as follows:
identical. The ranked order correlation between the two methods was K25 5 ~Cs/Cm!
unity. There was no direct correlation possible between numerical results (1)
obtained and reported by the two different methods. Abrasives that
showed qualifying extracted salts using the ISO Procedure also showed where:
qualifying extracted salt levels as specified in SSPC-AB 1. Cm 5 conductance, measured at 25°C (see 10.1), µmho,
Converting from ISO Reported Units to ASTM Reported Units:
Converting from one unit base to another is not useful as the two
and
methods differ in process. The conversion factor from µmho/cm to Cs 5 conductivity, 72 µmho/cm (from Table 1).
milliSiemens/m is as follows: NOTE 3—In general the cell constant is not greatly affected by
A Micro Mho Per Centimetre variations in the strength of the KCl solution, but, for greater accuracy,
1µmhocm–1 (1 3 10–6) V–1 cm–1 measurements should be made at or near the specific conductivity of the
A MilliSiemen Per Metre solution to be measured and at values that utilize the middle range of the
1 (MilliSiemen)(m–1) 5 (1310–3) V–1 100 cm–1 scale of the conductivity bridge, using the same multiplier tap.
Thus one milliSiemen/m 5 ten µmho/cm.
8.1.2 Table 1 gives values of specific conductivities for
5.2 Filter Paper, conforming to Specification E 832, Type corresponding KCl solution concentrations which are useful
1, Class C, to keep silt from fouling the surfaces of the for abrasive testing.
conductivity cell.
9. Procedure
6. Reagents and Materials
9.1 Preparation of a Slurry Filtrate:
6.1 Purity of Reagents—Reagent grade chemicals shall be 9.1.1 Rinse beakers, stirring rods, and funnels with reagent
used in all tests. Unless otherwise indicated, it is intended that water until tests show the rinse water has a conductivity of 5.0
all reagents conform to the specifications of the Committee on µmho/cm or less.
Analytical Reagents of the American Chemical Society where 9.1.2 Add 300 mL of water to 300 mL of abrasive and stir
such specifications are available.6 Other grades may be used, for 1 min with a stirring rod. Let stand for 8 min and then stir
provided it is first ascertained that the reagent is of sufficiently again for 1 min.
high purity to permit its use without lessening the accuracy of 9.1.3 Filter sufficient supernatant liquid for tests, discarding
the determination. the first 10 mL of the filtrate. The amount of supernatant liquid
6.2 Purity of Water—Unless otherwise indicated, references filtered shall be sufficient to cover the cell.
to water shall be understood to mean reagent water as defined 9.1.4 Rinse the conductivity cell in reagent water until the
by Type IV of Specification D 1193. rinse water is a cleanliness of 5.0 µmho/cm or less.
6.3 Potassium Chloride (KCl or 0.02 N KCl solution). 9.1.5 Rinse the conductivity cell two or three times with the
filtrate then determine conductance at 25°C in accordance with
7. Sampling
the operating instructions of the instrument. Use successive
7.1 Sampling shall be as follows unless otherwise agreed portions of the sample until a constant value is obtained.
upon between the purchaser and the seller. Take two 1-L
samples of abrasive at random from different packages of each 10. Calculation
lot, batch, day’s pack, or other unit of production in the 10.1 Calculate the specific conductivity of the abrasive as
shipment. When no markings distinguishing between units of follows:
production appear, take samples from the different packages in
Cs 5 C m 3 K25
the ratio of two samples for each 5000 kg, except that for (2)
shipments of less than 5000 kg, take two samples. Test the
samples separately.
TABLE 1 Specific Conductivities for Potassium Chloride (KCl)
Concentrations at 25°C
Heated, Dry KCl/Reagent KCl Conductivity,
6
Reagent Chemicals, American Chemical Society Specifications, American Normality
Water Solution, g/L µmho/cm
Chemical Society, Washington, DC. For suggestions on the testing of reagents not
listed by the American Chemical Society, see Analar Standards for Laboratory 0.0005 0.0373 72
Chemicals, BDH Ltd., Poole, Dorset, U.K., and the United States Pharmacopeia 0.001 0.0746 147
and National Formulary, U.S. Pharmacopeial Convention, Inc. (USPC), Rockville, 0.005 0.3728 718
0.01 0.7455 1414
MD.
2
D 4940
11. Report 12.1.1 Repeatability—Two results, each the mean of two
11.1 Report the following information: runs obtained by the same operator should be considered
11.1.1 The calibration value of the cell constant (both as suspect if they differ by more than 5 % relative.
measured and as predetermined and supplied with the 12.1.2 Reproducibility—Two results, each the mean of two
conductivity cell), the date, and the name of the person runs, obtained by operators in different laboratories should be
checking the calibration. considered suspect if they differ by more than 22 % relative.
11.1.2 The material, date, readings, and mean in µmho/cm 12.2 Bias:
along with name of person conducting the tests and 12.2.1 Bias can be present because of the mobility of
identification of the apparatus. various ions. The hydrogen ion has a much greater mobility
than the hydroxyl ion or other ions so that at low pH’s the
12. Precision and Bias 7
conductivity will be relatively higher than at high pH’s for the
12.1 Precision—On the basis of five replicate same ionic concentration. However, the bias introduced by this
interlaboratory tests of this test method in which three factor is in the proper direction. That is, high conductivity due
operators in three laboratories analyzed, in duplicate, six to a lower pH of the contamination would normally indicate
blasting abrasives containing ionogenic contamination, the greater corrosion potential.
within-laboratory coefficient of variation after rejecting results
12.2.2 A bias may be introduced by extraneous
from one set of replicate tests as outliers, was found to be
contamination or from reduced sensitivity of instruments for
1.7 % with 20 degrees of freedom (df) and the between-
low levels of contamination in the range of conductivity
laboratory standard deviation coefficient of variation was found
between 0 and 30 µmho/cm.
to be 7.4 % with 15 df. Based on these coefficients, the
following criteria should be used for judging the acceptability
13. Keywords
of results at the 95 % confidence level:
13.1 ionogenic; contamination; steel surfaces; abrasive;
blasting; conductimetric; analysis; interlaboratory testing;
7
Supporting data available from ASTM Headquarters. Request RR: D01-1061. precision; chloride; conductivity; salts.
The American Society for Testing and Materials takes no position respecting the validity of any patent rights asserted in connection
with any item mentioned in this standard. Users of this standard are expressly advised that determination of the validity of any such
patent rights, and the risk of infringement of such rights, are entirely their own responsibility.
This standard is subject to revision at any time by the responsible technical committee and must be reviewed every five years and
if not revised, either reapproved or withdrawn. Your comments are invited either for revision of this standard or for additional standards
and should be addressed to ASTM Headquarters. Your comments will receive careful consideration at a meeting of the responsible
technical committee, which you may attend. If you feel that your comments have not received a fair hearing you should make your
views known to the ASTM Committee on Standards, 100 Barr Harbor Drive, West Conshohocken, PA 19428.