Stopped-Flow Calorimeter For Biochemical Applications: Reagents
Stopped-Flow Calorimeter For Biochemical Applications: Reagents
Stopped-Flow Calorimeter For Biochemical Applications: Reagents
A rapid-response stopped-flow calorimeter for small samples of reagents is described. The construction,
performance characteristics and operational limitations are described, along with an example of its ability
to resolve the kinetics of an enzyme-catalysed hydrolysis. It is thought likely that the method would find
useful application in a variety of chemical and biochemical investigations.
Vol. 248
678 J. V. Howarth, N. C. Millar and H. Gutfreund
Pushing~~~~~~~~~~~~~~~~.
r~ ~ ~ ~ ~ ~ ~ ~ ~ ~ .
/~ ~ ~ ~ . . . . . .
syrnge ...............
Trigger
cells
Li
stat, and for the second part, by using for the reaction both directions of movement by needle valves restricting
chamber, a very-thin-walled tube insulated from the the exhaust side. The excursion of the main pushing
thermostat by an air jacket. This led to the configuration cylinder is restricted by two pieces of aluminium channel
as shown in Fig. 1. clipped over the through-rod at the front and rear ends.
The instrument described here is the one presently in The rear one carries a piezo-electric wafer, sealed into the
use and all of the experiments described in the accom- cut-offend of a plastic disposable syringe, which provides
panying paper (Millar et al., 1987) were made with this a trigger signal at the precise moment when the rod hits
instrument. the stop. Movement of the piston can be monitored by a
conducting plastic linear transducer that is also coupled
Features of the instrument to the rear extension of the through-rod but not shown
In use the whole assembly on its base-plate is immersed in the diagram.
in a well-stirred 80-litre water bath in a heavily insulated In this arrangement the flow is arrested upstream from
box. All operation of the instrument is from outside the the calorimeter, and after each push the delivery syringes
box by remote control. Except at 0 °C, when there is ice are re-charged with a single portion of reactant solution
present, no attempt is made to regulate the temperature (usually 150 1,) ready for the next push. More commonly
of the bath. Rather, it is adjusted initially and then in stopped-flow systems the flow is arrested by stopping
allowed to drift. In the course of a 4-5 h experiment the a collecting syringe on the downstream side of the
bath never drifts more than 0.1 K and then so slowly that observation cell. This method was not open to us because
the reference baseline for transient measurements is not it leads to a large pressure jump at the moment of stop,
affected. which would cause a large temperature change due to
Fig. 1 shows the instrument as it is normally used, with thermo-elastic heat released in the fluid (and more than
two calorimeters operated differentially. The valve block likely rupture the thin walls of the observation tube).
is made from aluminium alloy with a slot to take a close- Bowen et al. (1980) describe another solution to this
fitting polytetrafluoroethylene slider, which is drilled to problem using a push-pull flow arrangement. We have
form a four-ganged three-way tap so that in one position not tried this but it could be added to this instrument and
the delivery syringes are in straight-through communica- may be better than the simple upstream stop.
tion with the delivery tubes and thence to the calorimeter
cells, and in the other position the delivery syringes Calorimeter cells
connect to the storage syringes for re-charging. Actuation A calorimeter cell of the current type is illustrated in
of the slider is by the two air cylinders. The third air Fig. 2. It has four parts: a mixing chamber, a reaction
cylinder actuates the delivery syringes. All three air chamber, a thermopile and an air jacket.
cylinders are 25 mm-bore double-acting cylinders and
are supplied with air at 0.8 MPa (8 bar) pressure from Mixing chamber. The requirement of the mixing
outside the box via nylon tubes (which are not shown in chamber is that it provides for rapid and complete
the illustration). Speed of the cylinders is controlled in mixing but with the least amount of mechanical work,
1987
A stopped-flow calorimeter 679
IJ i~~~h
rc
mc
Fig. 2. Diagram of a calorimeter cell showing the mixing chamber and the reaction chamber in section
(a) A single calorimeter cell. Key: th, thermopile; mc, mixing chamber; rc, reaction chamber; dt, delivery tubes. The two
reactant solutions enter the mixing chamber at the sides and exit via the thin-walled tube that forms the reaction chamber. The
thickness of the thermopile is exaggerated in relation to the diameter of the reaction chamber. (b) Cross-section through the
thermopile at the level Y-Y. (c) Cross-section through the mixing chamber at the level X-X.
and hence heat, put into the fluid. We do not know of force of constantan/chromel couples is 59.1 uV * K-' at
any established theoretical treatment that would provide 273 K. Thermopiles made in this way give the full
a basis for such a design, so the form and dimensions of theoretical output provided that none of the couples is
the present chambers have evolved by intuition rather short-circuited. The thermopile is inserted through slits
than by design. Mixing chambers may be made by cut in the side of the tube, arranged so that the 'hot'
drilling in Perspex [poly(methyl methacrylate)]. This is junctions lie along the midline, and is then sealed in with
satisfactory for single cells, but for paired cells to be used a small amount of epoxy resin.
in a differential mode it is important to have the two
chambers exactly alike and it is difficult to machine them Air jacket. The air jacket is simply the end cut from a
with the necessary precision. The two mixing chambers 5 ml plastic disposable syringe. The thermopile passes
in the present cells were made by casting liquid epoxy through, and is sealed into, slits cut in the wall so that the
resin, both in the same mould. This method has given the electrically insulated 'cold' junctions protrude into the
best results to date. It is important to ensure that the bath water.
reaction chamber tube is truly central where it projects
into the mixing chamber and that the gap between the Characteristics of the instrument
end of the tube and the back of the mixing chamber is Three characteristics determine the usefulness of the
exactly the same in the two cells. It is our practice to view instrument in a particular application. They are sensi-
the mixing chambers under 20 x magnification and to tivity, speed of response and rate of heat loss.
remove surface irregularities at this level.
Sensitivity. Fig. 3(a) shows a series of records obtained
Reaction chamber. The reaction chamber, where the by reaction of different concentrations of HCl with Tris
measurements are made, is a tube that carries the buffer at pH 8. In Fig. 3(b) the output voltage is plotted
reaction solution from the mixing chamber. The require- against temperature change calculated from the known
ments are that the heat capacity of the wall must be small enthalpy change of the reaction. The sensitivity is given
compared with that of the contained fluid and that the by the slope of the regression line, giving a value of
wall must come rapidly into temperature equilibrium 5.95 mV K-', exactly the same as the expected theoretical
-
with the fluid. Both of these are met by having a thin- value for 100 thermojunctions. This procedure gives the
waIled tube. Our tubes are made by spiral wrapping a true temperature sensitivity of the instrument, but
tape, cut from 10,tm polyester film, around a 2 mm (unlike a batch calorimeter) it slightly overestimates the
polythene former, the width of the tape and the pitch of heat calibration because it does not take account of
the spiral being calculated to give a double thickness of the heat capacity of the thermopile and the wall of the
film at all points along the wall. The tape is coated with reaction chamber. This is because the neutralization
epoxy resin as it is wound and excess resin is removed. reaction is so quick that all the heat is released before the
When the resin has hardened the polythene former is flowing solution reaches the thermopile. Typically the
withdrawn. These tubes weigh 2.37 mg for 5 mm length flow continues for 30 ms or more, giving plenty of time
(i.e. the length of the thermopile) and have a heat for the thermopile and wall to come into temperature
capacity of about 2.76 mJ * K-1, or about 4 % that of the equilibrium. Thus, when the flow finally stops, the fluid
contained solution. in contact with the thermopile does not lose any heat to
these 'inert' elements. On the other hand, the enzyme
Thermopile. The thermopiles are made from 25 ,m- reactions that the instrument is designed to investigate
diameter constantan and chromel wire. The wires are are not so fast and have scarcely got underway before the
embedded in epoxy resin and insulated by sandwiching solution reaches the thermopile after mixing. Thus, when
between two layers of 10 ,um polyester film. The the flow is arrested, all subsequent reaction heat is shared
thermopile consists of 100 couples in series, has an with the inert elements, leading, in the present instrument,
electrical resistance of 2300 Q and extends 5 mm along to a 5 0 reduction in temperature change.
the line of flow. The heat capacity of the thermopile in
contact with the solution is 0.8 mJ K-1, which is close to
-
Speed of response. Fig. 4 shows a record obtained by
1 % of that of the solution. The thermo-electromotive passing a 2 ms pulse of current between electrodes placed
Vol. 248
680 J. V. Howarth, N. C. Millar and H. Gutfreund
61o (a)
41
20
0 -
4(U
Q
-1.0 -0.5 0 0.5 1.0 0 10 20 30
Time (ms)
I-
Time (s)
Fig. 4. Response time of the calorimeter
L..
Q
A 2 ms pulse of heat (marked by the horizontal bar) was
60 1_ (b) produced by passing an electric current pulse through an
electrolyte solution in the reaction chamber. This is the
average of six records with the pulse polarity reversed
40 _ 00-0 between successive records. Time zero marks the time at
which the pulse starts. The fitted exponential is fitted from
2 ms and has a rate constant of 302 s-'.
20 0
0 1 1
which flow stops. (b) Peak thermopile output versus HCI .91
1987