Manual-1515674466-Sh marLED E9 E15 10 en 01 K
Manual-1515674466-Sh marLED E9 E15 10 en 01 K
Manual-1515674466-Sh marLED E9 E15 10 en 01 K
Produced By:
Alcon Laboratories, Inc.
15800 Alton Parkway
Irvine, California 92618-3818
U.S.A.
Telephone: 949/753-1393
800/832-7827
FAX: 949/753-6614
ii 8065751131
TABLE OF CONTENTS
TOPIC PAGE #
Manual Revision Record . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ii
Foreword . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . vi
Important Notice. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . vii
SECTION ONE - GENERAL INFORMATION
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1.1
Technical Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1.2
Laser Characteristics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1.2
Environmental Considerations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1.3
Universal Precautions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1.4
EMC Statement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1.4
FCC and IC Compliance Statement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1.7
Labeling . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1.8
Carring Case . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1.9
Preparing For Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1.10
PurePoint™ Laser Safety Features . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1.13
Professional Operator’s Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1.14
Product Service. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1.17
Limited Warranty . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1.18
SECTION TWO - DESCRIPTION
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2.1
Front Panel Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2.1
Rear Panel Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2.7
Footswitch . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2.8
System Modes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2.9
Treatment Modes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2.9
Screen Descriptions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2.10
SECTION THREE - OPERATING INSTRUCTIONS
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3.1
1 Initial Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3.1
2 System Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3.1
3 System Power Up And Set Up . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3.4
4 Normal Operating Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3.7
5 Turn Off Sequence . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3.9
6 Changing The System Settings. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3.10
7 Identifying Unrecognized Probes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3.11
SECTION FOUR - CARE AND MAINTENANCE
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4.1
Care And Cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4.1
Fuse Replacement Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4.2
Calibration Verification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4.3
System Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4.9
Aiming Beam/LIO Illumination Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4.11
SECTION FIVE - TROUBLESHOOTING
System Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5.1
SECTION SIX - ACCESSORIES AND PARTS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6.1
SECTION SEVEN - INDEX . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7.1
8065751131 iii
LIST OF ILLUSTRATIONS
FIGURE # TITLE PAGE #
Figure 1-1 The Alcon PurePoint™ Laser . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1.1
Figure 1-2 Labels and Icons used on the PurePoint™ Laser Console . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1.7
Figure 1-3 The PurePoint™ Carrying Case . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1.8
Figure 1-4 Recommended Laser Room Layout . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1.10
Figure 1-5 Remote Connector/Door Lamp Circuit Diagram . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1.12
Figure 1-6 Dr. Filter Message . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1.15
Figure 2-1 The PurePoint Laser Front Panel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2.1
Figure 2-2 Typical PurePoint™ Screen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2.2
Figure 2-3 Unidentifiable Probe Selection Screen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2.3
Figure 2-4 The PurePoint™ Laser Rear Panel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2.2
Figure 2-5 The PurePoint™ Laser Footswitch. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2.7
Figure 2-6 Screens Displayed on the PurePoint™ LCD . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2.10
Figure 2-7 Initialization Screen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2.11
Figure 2-8 Standby Screen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2.11
Figure 2-9 Ready Screen. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2.12
Figure 2-10 Laser On Screen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2.12
Figure 2-11 Main Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2.12
Figure 2-12 Settings Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2.13
Figure 2-13 Audio Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2.13
Figure 2-14 Contrast Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2.14
Figure 2-15 Fiber Custom Pre-Sets Screen. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2.14
Figure 2-16 Footswitch Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2.15
Figure 2-17 Language Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2.15
Figure 2-18 Revert to Standby Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2.16
Figure 2-19 Treatment Totals Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2.16
Figure 2-20 System Information Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2.17
Figure 2-21 About PurePoint™ Screen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2.17
Figure 2-22 Version Numbers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2.17
Figure 2-23 System Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2.18
Figure 2-24 System Totals . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2.18
Figure 2-25 Reset to Factory Defaults Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2.18
Figure 3-1 Slit Lamp Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3.3
Figure 3-2 Display During Initialization . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3.4
Figure 3-3 Transition from Standby to Ready Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3.8
Figure 3-4 Transition from Ready State to Laser On . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3.8
Figure 3-5 Transition from Laser On to Standby because of a Warning Condition . . . . . . . . . . . . . . . . . . 3.9
Figure 3-6 Main Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3.10
Figure 3-7 Settings Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3.10
Figure 3-8 Unidentifiable Probe Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3.11
Figure 4-1 Exposure Time Test Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4.4
Figure 4-2 Power Test Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4.5
Figure 5-1 Error Message . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5.2
Figure 6-1 Alcon SL1000 Slit Lamp . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6.2
Figure 6-2 Zeiss 30SL. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6.2
Figure 6-3 Haag-Streit 900 BM . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6.3
Figure 6-4 Doctor Protection Filter Mounted Between Binoculars and Slit Lamp Assembly . . . . . . . . . . . 6.4
Figure 6-5 Label Location Diagram on Adaptation - Alcon SL 1000 shown . . . . . . . . . . . . . . . . . . . . . . . . 6.6
Figure 6-6 To avoid injury, the beam splitter/accessories must be placed between the binoculars
and Doctor Protection Filter (Alcon SL 1000 shown) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6.6
Figure 6-7 Controls on Alcon SL 1000 Slit Lamp with Doctor Protection Filter and Adaptation Installed. 6.7
Figure 6-8 Laser Spot Focus . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6.9
iv 8065751131
LIST OF ILLUSTRATIONS
FIGURE # TITLE PAGE #
Figure 6-9 Alcon LIO-AT . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6.10
Figure 6-10 Alcon LIO-AT Labeling . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6.12
Figure 6-11 Adjusting the LIO-AT Overband . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6.14
Figure 6-12 LIO-AT Controls and Adjustments . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6.15
Figure 6-13 Eyecup Retainers and Ocular Lens on the Alcon LIO-AT . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6.16
Figure 6-14 Alcon LIO-AT Bulb Replacement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6.20
LIST OF TABLES
TABLE # TITLE PAGE #
8065751131 v
FOREWORD
This Operator's Manual is designed to acquaint the operator and operating room personnel
with the Next Generation Laser. The manual presents an organized summary of the operating
principles, main components, safety features, and instructions for care and use of the
instrument.
The information in this manual should be supplemented with reference works on laser theory
and the interaction of laser energy with biologic tissues. No attempt is made in this manual
to answer all the questions that arise during the use of the instrument in medical procedures.
Please contact Alcon for complete technical support and service if you have questions
concerning any aspect of this instrument's operation or if it fails to perform satisfactorily.
vi 8065751131
IMPORTANT NOTICE
Equipment improvement is an on-going process and, as such, changes may be made to the
equipment after this manual is printed.
Pay close attention to WARNINGS and CAUTIONS in this manual. WARNINGS are written
to protect individuals from bodily harm. CAUTIONS are written to protect the instrument
from damage. Illustrations contained in this manual are for reference only.
Alcon Surgical shall not be liable for any damage resulting from failure to comply with the
enclosed instructions.
CAUTION
U.S. Federal Law restricts this device to sale by or on the order of a physician only.
WARNINGS!
Use of controls or adjustments, or performance of procedures other than those specified
herein may result in hazardous laser radiation exposure.
A qualified technician must check ground continuity and both polarities for leakage
current every twelve months to ensure they are within the applicable standard (for
example: EN 60601-1/IEC 601-1). Values must be recorded, and if they are above the
applicable standard, or 50% above your first measurement, do not use the system; call
Alcon Technical Services.
WARNING!
Use of accessories and cables other than those provided may result in increased
emissions or decreased immunity of the system. Portable and mobile RF
communications equipment can affect this medical electrical equipment.
All rights reserved. No part of this manual may be reproduced, transmitted, or stored in
a retrieval system, in any form or by any means; photocopying, electronic, mechanical,
recording, or otherwise; without prior written permission from Alcon Laboratories, Inc.
8065751131 vii
THIS PAGE INTENTIONALLY BLANK
INTRODUCTION
The system is also supported by a wide range of high quality laser probes, a laser
indirect ophthalmoscope (LIO), and is compatible with a wide variety of slit lamps.
8065751131 1.1
Table 1-1
Technical Specifications
CATEGORY SPECIFICATION
Table 1-2
Laser Characteristics
CATEGORY TREATMENT LASER BEAM AIMING LASER BEAM
Laser Class IV II
Laser Power • 30 mW to 200 mW in 10 mW Less than 1 mW; adjustable by
steps operator
• 200 mW to 500 mW in 20 mW
steps with additional steps at:
250, 350, 450
• 500 mW to 950 mW in 50 mW
steps
1.2 8065751131
WARNINGS!
There are potential hazards when inserting, steeply bending, or improperly securing
the fiber optic. Not following the recommendations of the manufacturer may lead
to damage to the fiber or delivery system and/or harm to the patient or user.
Since the aiming beam passes down the same delivery system as the treatment
beam, it provides a good method of checking the integrity of the delivery system.
If the aiming beam spot is not present at the distal end of the delivery system,
or its intensity is reduced or it looks diffused, this a possible indication of a
damaged or not properly working delivery system. If there is any doubt, contact
Alcon Technical Services.
The use of flammable anesthetics or oxidizing gases such as nitrous oxide (N2O)
and oxygen should be avoided. Some materials - for example cotton wool when
saturated with oxygen - may be ignited by the high temperatures produced in normal
use of the laser equipment. The solvents of adhesives and flammable solutions
used for cleaning and disinfecting should be allowed to evaporate before the laser
equipment is used. There is also danger of ignition of endogenous gases.
NOTE: To eliminate power consumed when the key switch is off, turn off the
main power switch on the rear panel.
Environmental Considerations
The equipment that you have purchased requires the use of natural resources for its
production. This equipment may also contain hazardous substances which could have
potential effect on the environment and human health if disposed of improperly.
In order to avoid the entry of any such substances into our environment and to
promote natural resource conservation, we encourage you to use the appropriate take-
back systems. Such take-back systems reuse or recycle many of the materials in your
end-of-life equipment in a beneficial way. Please contact your local Alcon office for
assistance in take-back options through Alcon or other providers.
The crossed-bin symbol located on this equipment reminds you to use take-back
systems, while also emphasizing the requirement to collect waste equipment
separately, and not dispose of it as unsorted municipal waste.
If you need more information on the collection, reuse, or recycle systems available to
you, please contact your local or regional waste administration, or contact your local
Alcon office for more information.
8065751131 1.3
Universal Precautions
Universal precautions shall be observed by all people who come in contact with the
instrument and/or accessories to help prevent their exposure to blood-borne pathogens
and/or other potentially infectious materials. In any circumstance, wherein the exact
status of blood or body fluids/tissues encountered are unknown, it shall be uniformly
considered potentially infectious and handled in accordance with OSHA guidelines.
EMC Statement
It is important to install and use the equipment in accordance with the instructions
in order to prevent harmful interference with other devices in the vicinity. If this
equipment causes harmful interference to other devices (determined by turning the
equipment off and on), the user is encouraged to try to correct the interference by one
or more of the following measures:
Voltage fluctuations/ Complies The EMC Statement provides guidance on steps to take in case of electromagnetic
Flicker emissions interference.
IEC 61000-3-3
1.4 8065751131
Table 1-4 Guidance and Manufacturer's Declaration - Electromagnetic Immunity - The
PurePoint™ Laser is intended for use in the electromagnetic environment
specified below. The customer or the user of the Next Generation Laser should
assure that it is used in such an environment.
Immunity Test IEC 60601 Test Level Compliance Level Electromagnetic Environment-Guidance
Electrostatic • ±6 kV contact • ±6 kV contact Floors should be wood, concrete, or ceramic tile.
discharge (ESD) • ±8 kV air • ±8 kV air Do not use around floors that are covered with
IEC 61000-4-2 synthetic material to avoid laser stoppage due to
ESD.
Electrical fast • ±2 kV for power • ±2 kV for power Mains power quality should be that of a typical
transient/burst supply lines supply lines commercial or hospital environment. To avoid laser
IEC 61000-4-4 • ±1 kV for input/output • ±1 kV for input/ stoppage due to fast transients avoid powering the
lines output lines PurePoint™ Laser on the same branch circuit with
sources that can generate fast transients (inductive
switching; e.g., high current motors).
Surge • ±1 kV differential • ±1 kV differential Mains power quality should be that of a typical
IEC 61000-4-5 mode mode commercial or hospital environment. To avoid
• ±2 kV common mode • ±2 kV common laser stoppage due to power-line surges consider
mode powering the PurePoint™ Laser through branch
circuit that has surge suppressor for protection
against lightning surges (e.g., at power panel to
surgical/office suite).
Voltage dips, short • <5% UT (>95% dip in • <5% UT (>95% dip Mains power quality should be that of a typical
interruptions, and UT) for 0.5 cycle in UT) for 0.5 cycle commercial or hospital environment. If the uses of
voltage variations • 40% UT (60% dip in • 40% UT (60% dip the PurePoint™ Laser require continued operation
on power supply UT) for 5 cycles in UT) for 5 cycles during power mains interruptions, it is recommended
input lines • 70% (30% dip in UT) • 70% (30% dip in that the PurePoint™ Laser be powered from an
IEC 61000-4-11 for 25 cycles UT) for 25 cycles uninterruptible power supply or a battery.
• <5% (>95% dip in • <5% (>95% dip in
UT) for 5 sec UT) for 5 sec
Power frequency 3 A/m 3 A/m Power frequency magnetic fields should be at
(50/60 Hz) levels characteristic of a typical location in a typical
magnetic field commercial or hospital environment.
IEC 61000-4-8
Conducted RF 3 Vrms 3 Vrms Portable and mobile RF communications equipment
IEC 61000-4-6 150 kHz to 80 MHz should be used no closer to any part of the
PurePoint™ Laser, including cables, than the
recommended separation distance calculated
from the equation applicable to the frequency to
the transmitter.
8065751131 1.5
Table 1-5 Recommended Separation Distances Between Portable and Mobile RF Communi-
cations Equipment and the PurePoint™ Laser - The PurePoint™ Laser is intended
for use in an electromagnetic environment in which radiated RF disturbances are
controlled. The customer or the user of the PurePoint™ Laser can help prevent
electromagnetic interference by maintaining a minimum distance between porta-
ble and mobile RF communications equipment (transmitters) and the PurePoint™
Laser as recommended below, according to the maximum output power of the
communications equipment.
Separation distance according to frequency of transmitter
(m)
Rated maximum output 150 kHz to 80 MHz 80 MHz to 800 MHz 800 MHz to 2.5 GHz
power of transmitter d = 1.2√P d = 1.2√P d = 2.3√P
(W)
0.01 0.12 0.12 0.23
0.1 0.38 0.38 0.73
1 1.2 1.2 2.3
10 3.8 3.8 7.3
100 12 12 23
For transmitters rates at a maximum output power not listed above, the recommended separation distance d in meters (m) can
be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the
transmitter in watts (W) according to the transmitter manufacturer.
Note 1 - At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
Note 2 - These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from
structures, objects, and people.
1.6 8065751131
FCC and IC Compliance Statement
This device complies with Part 15 of the FCC Rules and with Industry Canada Radio
Standards Specification RS-210.
CAUTIONS
Change or modifications made to this equipment not expressly approved by Alcon
may void the FCC authorization to operate this equipment.
To ensure that the RFID transmitter complies with current FCC regulations limiting
both maximum output RF power and human exposure to radio frequency radiation, a
separate distance of at least 20 cm must be maintained between the unit’s antenna and
the body of the user and any nearby persons at all times and in all applications and
uses.
This device complies with the RF exposure limits for humans as called out in
RSS-102.
This device complies with the requirements of the Council Directive 99/5/EC
(R&TTE).
This device complies with the Australian/New Zealand Standard AS/NZS 4268:
2003 Radio Equipment and Systems – Short Range Devices – Limits and methods
measurement.
Canada
This ISM device complies with Canadian ICES-001.
(Cet appareil ISM est conforme a la norme NMB-001 du Canada.)
8065751131 1.7
LABELING
Aiming Beam On
!
Dangerous Voltage Consult Operator's Manual, or
System Error or Advisory
Equipotentiality
Ready
Footswitch
Standby State
Fuse
SYSTEM
FAULT
System Fault
Illumination
System Information
Keyswitch
Type BF Equipment
Laser Connection
USB Connector
1.8 8065751131
Carrying Case
The carrying case shown in Figure 1-4 is
included with the system and intended to be
used as an aid to carrying the system.
CAUTION
The carrying case should not be used for
shipping the system.
8065751131 1.9
PREPARING FOR INSTALLATION
The PurePoint™ Laser system was thoroughly inspected and carefully packaged
for shipping. If the container is damaged, leave system in original container with
packaging and request inspection by the carrier within 3 days of delivery.
Included as part of the packaging is the carry box for the PurePoint™ Laser. This
container is intended to protect the system when moving it from one location to
another. Use the carry box whenever the system must be moved.
1.10 8065751131
NOTE: The accessory equipment connected to or used with this equipment must
be certified according to the respective IEC standard; e.g., IEC 950 for
data processing equipment (data processing equipment must not be
used during patient treatment) and IEC 601-1 for medical equipment.
Additionally, all configurations shall comply with the system standard
IEC 601-1-1. Anyone connecting additional equipment or otherwise
causing a different system configuration than provided by Alcon, is
responsible for continued compliance to the requirements of the System
Standard IEC 601-1-1. If in doubt, consult the Technical Services
department of your local Alcon representative.
8065751131 1.11
REMOTE
CONNECT
1
CUSTOMER SUPPLIED
DOOR SWITCH (OPTIONAL)
6
9-PIN D-CONNECTOR
1
COM
TO DRIVE DOOR LAMP (OPTIONAL)
NC NO 9
3
Utility Requirements
Electrical requirement: The PurePoint™ Laser has a power supply that operates at
100-120 V and 220-240 V input ranges at 50/60 Hz. A properly grounded, standard
plug is the only requirement.
Electrical Connections
CAUTION
Before turning the instrument ON for the first time after receipt of the system, wait one
hour for the components and optics to normalize to avoid possible condensation that
may have occurred during shipping.
Optical Connections
Optical connections vary in relation to the procedure to be performed. Different
peripherals can be connected to the output ports. These peripherals are:
• Slit Lamp adaptation
• Laser Indirect Ophthalmoscope (LIO)
• Endoprobe/Aspirating Endoprobe/Illuminated Endoprobe
The procedures for connecting these peripherals are contained in Section Three:
Operating Instructions.
1.12 8065751131
PUREPOINT™ LASER SAFETY FEATURES
The PurePoint™ Laser is designed for the highest degree of reliability and maximum
safety for both the operator and the patient. Any misuse of this laser system may
be dangerous. Before using the laser system, the operator must be familiar with the
commands and the manipulation of this type of instrument.
The PurePoint™ Laser is fitted with the following safety systems which must be
understood by every operator:
• A protective housing covers the laser source so that no harmful laser radiation will
be emitted. No part of this protective housing should be removed by the operator.
The laser system must not be used if the protective housing has been damaged or
removed.
• A remote connection (interlock) is located on the rear panel and permits
the installation of an external switch. Refer to Figure 1-4 for remote switch
connections. This switch can be installed on the laser room door and cuts off all
laser emissions in case the door is opened during operation. There is also a relay
connector for connection to an internal relay to activate a door warning lamp if
desired.
• A key switch controls the laser power supply. Laser operation is not possible if
the key has been removed. Access to the key should be limited to authorized and
knowledgeable personnel. The key should not be left on or near the instrument
when not in use.
• During operation, laser status can be determined by visually checking the LCD
display. The background colors change to indicate the laser’s status for Standby or
Ready modes.
• A green background display indicates that the system is in Ready mode. In
addition, the system will emit a tone to indicate the mode change. The power and
time settings can be set or changed, but not while the laser is being fired.
• A gray background on the LCD indicates that the system is in Standby mode and
the laser’s default parameters can be altered.
• Under normal Standby, non-firing situations, the background display will be grey.
• An emergency switch is mounted on the front panel. Pushing this switch will cut
off all laser emissions (treatment and aiming beam) at any time. The switch must
be pulled out to the initial position to restore power. The laser will always restart
in Standby mode.
• Laser firing commands are microprocessor controlled and firings are prevented
should any malfunction be detected in the instrument electronics. The instrument
will only fire when all conditions are correct.
• Output power of the laser beam is continuously monitored and controlled. In
case an unusual power condition is detected, firing stops and the treatment laser
emission is cut off.
8065751131 1.13
PROFESSIONAL OPERATOR’S INFORMATION
The following information is given to provide the operator with specific information
regarding the PurePoint™ Laser ophthalmic laser.
Indications
The PurePoint™ Laser is indicated for use in photocoagulation of both anterior and
posterior segments of the eye including:
1.14 8065751131
Effects
The laser beam is primarily absorbed by pigmented tissues within the eye. These
primary pigments are hemoglobin/oxy-hemoglobin and melanin. In the case of
macular treatment, xanthophyll pigment is involved. The surgeon controls the power,
spot size, and exposure time of the delivered laser beam to the targeted tissue. It is
the combination of these effects that results in the thermal action of the laser beam
upon tissue. One or all of the adjustable parameters can be changed. However, in
normal clinical practice, power is usually varied, and spot size and exposure time are
preset as a function of the application.
The 532 nm green laser beam has similar absorption characteristics to the 577 nm dye
yellow laser beam7. This means that the absorption effects of the 532 nm wavelength
are considerably higher in hemoglobin and melanin, and less in xanthophyll. In
all cases, it is necessary to perform titrations until the desired treatment results are
obtained. The 532 wavelength also requires less power than that required with the
argon laser to obtain similar results. Therefore, you should begin your titration levels
with lower power than required for similar procedures with the argon laser.
WARNING!
Failure to titrate delivered energy may result in patient injury.
Use of this medical laser, as with any other instrument, requires training and
experience to obtain maximum clinical performance. Titrating the dosage is
recommended by initiating a lesion formation in an area of normal retina with intact
pigment epithelium. Power and exposure duration should be varied incrementally
until the desired lesion is produced.
WARNING!
If unsure which settings are required, select low power, short duration, and large
spot size. Failure to do so may result in patient injury.
8065751131 1.15
Delivery of Laser Energy
The laser beam is delivered to tissue via a Slit Lamp, Endoprobe, Illuminated
Endoprobe, aspirating Endoprobe or Laser Indirect Ophthalmoscope (LIO). When
using a Slit Lamp, the laser beam is often used in combination with various contact
lenses to aid in treatment of particular targets such as the fundus. These contact lenses
enable the laser beam to be directed to different sections of the eye.
WARNING!
Some contact lenses, generally classified as wide field or pan fundus lenses, magnify
the laser spot incident upon them. For example, a pan retinal photocoagulation
procedure is normally done with a spot size setting of 500 microns when using
a three mirror lens. If a wide field lens were used, and the laser spot size setting
remained at 500 microns, the actual spot in the eye would be larger than the
indicated spot size setting. Normal increases in spot size in the eye range between
1.3 and 2 times the spot size as selected at the Slit Lamp zoom. These effects
and resultant changes in power density must be considered when using wide
field lenses.
Reaction to applied laser energy by the eye is a function of many variables. The
pigmentation of the eye, technique or procedure used, laser settings, and pre-existing
condition of the eye, such as cataract, will have an effect on the selected laser
parameters and the results obtained. Therefore, it is very important to consider all the
existing clinical conditions and titrate until the desired results are obtained.
Always use minimal illumination from the Slit Lamp while maintaining good
visualization in order to reduce reflections and discomfort for the patient. Likewise,
the aiming beam should be used at a minimum setting while maintaining proper
targeting of the selected tissue. This will also minimize excessive reflections and
scattering, particularly at smaller spot sizes.
The Doctor Protection Filter must remain in the beam path during treatment, enabling
the targeted tissue to be seen with complete protection for the operator. The filter has
virtually no effect on visualization (colored** view only).
Rotation of the tethered filter with manual switch in or out of the beam path is
accomplished by means of a lever located on the right side of the filter. Note that if
the Doctor Protection Filter is in the open position in the Slit Lamp or Endo modes,
the laser will not fire and the message “Please Engage Dr. Filter” will appear. Rotate
the filter lever clockwise until the Doctor Protection Filter is in the beam path and the
1.16 8065751131
message clears. If using a non-tethered fixed filter, and
the system is switched from Standby to Ready mode,
the message "Verify appropriate Dr. Filters are installed
in all viewing devices" appears and the user must verify
before the laser can switch to Ready mode.
WARNING!
Do not attempt treatment if aiming beam is not present. Patient injury may
occur.
NOTE: The aiming beam passes down the same delivery system as the working
beam; this provides a good method of checking the integrity of the
delivery system. If the aiming beam spot is not present at the distal end
of the delivery system, if its intensity is reduced, or if it looks diffused
these are possible indications of a damaged or not properly working
delivery system.
Treatment Hazards
A single treatment exposure will typically cause a blanching of target tissue. Exposure
duration can be adjusted from 0.01 seconds to 2.0 seconds to result in the desired
effect. A continuous treatment beam can also be selected.
NOTE: In CW, depending on the thermal load of the system, the system may
shut down in safety mode prior to the footswitch being released.
Contra Indications
Patients with a condition that prevents visualization of target tissue (cloudy cornea,
or extreme haze of the aqueous humor of the anterior chamber or vitreous humor) are
poor candidates for Slit Lamp or LIO delivered laser treatment.
8065751131 1.17
Side Effects
Corneal burns, inflammation, loss of best-corrected visual acuity, loss of visual field,
and transient elevations in intraocular pressure can occur as a result of ophthalmic
laser treatment. Unintentional retinal burns can occur if excessive treatment beam
power or duration is used.
Laser Safety
Back scattered radiation is of low intensity and is not harmful when viewed through
a protective filter. All personnel in the treatment room must wear protective eyewear,
OD 4 or above at 532 nm, when the system is in Standby/Ready mode as well as
during treatment. The Doctor Protection Filter is an OD greater than 4 at 532 nm.
WARNING!
Use of controls or adjustments or performance of procedures other than those
specified herein, may result in hazardous laser radiation exposure.
CAUTION
Federal (USA) law restricts this device to sale by, or on the order of, a physician.
1 Ludwig, K.; Lasser, T.; Sakowski, H.; Abramwoski, H.; Worz, G. (Augenklinik,
Universitat Munchen) "Photocoagulation in the edematous and non-edematous
retina with the cw-laser of different wavelengths." Ophthalmologe (GERMANY),
December 1994, Volume 91, No. 6, p783-788.
2 Roider, J.; Schiller, M.; el Hifnawi, E.S.; Birngruber, R. (Augenklinik,
Medizinische Universitat zu Lubeck) "Retinal photocoagulation with a pulsed,
frequency-doubled Nd: YAG laser (532 nm)." Ophthalmologe (GERMANY),
December 1994, Vol. 91 No. 6, p777-782.
3 Wyman, D.; Wilson, B.; Adams, K. (Medical Physics Department, Hamilton
Regional Cancer Centre, Ontario, Canada) "Dependence of laser photocoagulation
on interstitial delivery parameters." Lasers Surgical Medical (UNITED
STATES), 1994, Vol. 14 No. 1, p59-64.
4 Obana, A.; Miki, T.; Matsumoto, M.; Ohtsuka, H.; Moriwaki, M.; Kamo, M.;
Mii, T.; Kijima, M. (Department of Ophthalmology, Osaka City University,
Medical School, Japan) "An experimental and clinical study of chorioretinal
photocoagulation using a frequency-doubled Nd: YAG laser." Nippon Ganka
Gakkai Zasshi (JAPAN), September 1993, Vol. 97 No. 9, p1040-1046.
5 Mordon, S.; Beacco, C.; Rotteleur, G.; Brunetaud, J.M. (INSERM - National
Institute of Health and Medical Research - Lille, France) "Relation between skin
surface temperature and minimal blanching during argon, Nd-YAG 532, and CW
dye 585 laser therapy of port-wine stains." Lasers Surg Med (UNITED STATES)
1993, Vol. 13 No. 1, p124-126.
6 Jalkh, A.E.; Pflibsen, K.; Pomerantzeff, O.; Trempe, C.L.; Schepens, C.L. (Eye
Research Institute of Retina Foundation, Boston, MA 02114) "A new solid-
state, frequency-doubled neodymium-YAG photocoagulation system." Arch
Ophthalmol (UNITED STATES) June 1988, Vol. 106 No. 6, p 847-849.
7 Wavelengths, Opthamology, July 1986, Volume 93, Number 7, Page 956.
1.18 8065751131
PRODUCT SERVICE
For product service, please contact Alcon’s Technical Services Department at the
number provided below.
Operators experiencing problems with the system should refer to the Operating
Instructions and Troubleshooting sections of this manual. A problem which persists
should be referred to the Alcon Technical Services Department or your local
authorized service representative.
Safety performance should be verified by the user (e.g., qualified service personnel) at
least twice a year. Ground resistance must be under 0.1 ohms. Leakage current must
be under 500 µA.
8065751131 1.19
LIMITED WARRANTY
Alcon Laboratories, Inc., will repair or replace at its option, any system or
accompanying accessories (excluding the optical fiber) found to be defective in
material and/or workmanship for a period of one (1) year from the date of initial
installation. This warranty applies to the original purchaser of the system, when said
system is properly installed, maintained, and operated in accordance with published
instructions.
Alcon Laboratories shall not be obligated to provide services under this warranty for
damage to or destruction of systems covered where such damage or destruction is (i)
a result of or caused by fire or explosion of any origin, riot, civil commotion, aircraft,
war, or any Act of God including, but not limited to lightning, windstorm, hail,
flood, earthquake, or (ii) caused by customer’s misuse or improper servicing of said
systems.
The express warranty above is the sole warranty obligation of Alcon, and the
remedy provided above is in lieu of any and all other remedies. There are no other
agreements, guarantees, or warranties, oral or written, express or implied, including
without limitation warranty of merchantability or fitness for a particular purpose.
Alcon shall have no liability whatsoever for any incidental or consequential damages
arising out of any defect, improper use, or unauthorized service or repair.
WARNING!
The disposables used in conjunction with Alcon instrument products constitute
a complete surgical system. Use of disposables and handpieces other than those
manufactured by Alcon may affect system performance and create potential
hazards. If it is determined that disposables or handpieces not manufactured
by Alcon have contributed to the malfunction of the equipment during warranty
period, service will be provided at prevailing hourly rates.