ICPractical Ion Chromatography
ICPractical Ion Chromatography
ICPractical Ion Chromatography
Chromatography
An Introduction
Practical Ion Chromatography
An Introduction
4 Metrohm Monograph
The authors
Claudia Eith
Studied chemistry at the Fachhochschule in Aalen, one term's practical work in the drinking water and
wastewater analysis sector in Adelaide (Australia); since 2000 with Metrohm AG in the R&D division.
Maximilian Kolb
Studied chemistry at the Technical University in Munich, promotion in the field of
homogeneous catalysis, then managed the water quality division of the water authority in
Traunstein for 5 years. Since 1982 Professor at the Fachhochschule Aalen; fields of
work: environmental technology, environmental analysis and chemometrics.
Andreas Seubert
Studied chemistry at Hannover University, promotion 1990: "Ultratrace analysis in highly
pure refractory metals with trace-matrix separation by ion chromatography". Habilitation
1995: "On-line HPLC-atomic spectrometry coupling applications in elemental analysis",
1998 to 2000 temporary professorship in ana-lytical chemistry at Kassel University, since
March 2000 professorship in analytical chemistry at the Philipps University in Marburg.
Examining things which do not reveal themselves directly is always a challenge. The reasons for this
vary from simple curiosity to the real necessity of survival. There are many different ways of looking
behind the curtains. The simplest way is to use the human senses: listening, touching, smelling,
tasting and seeing. In earlier days the alchemists liked to use these five senses. This is why today
acids taste acidic and bromine has its name derived from "bromos", Greek for fetid. To the naked
eye chromium appears to be colored, as "chroma" from a historical linguistic point of view is the
same as color. The alchemists also expressed their feelings with insults such as "you kobold" for
cobalt, the presence of which caused our ancestors great difficulty in the production of iron.
Many individual things cannot be seen directly. They are too thoroughly mixed or the
human senses are not capable of identifying them. This is the moment when analysis
comes into play. It is able to extract precise information from an undefined mixture of
components which cannot be obtained by the human senses alone.
Although the organism is full of them, the human senses cannot experience them directly. We
are talking about ions, those charged atoms or molecules which are an integral part of virtually
all living and dead matter. Ions are responsible for the transfer of information along the nerves,
for ensuring that digestion takes place, for making sure that the blood pressure is correct and
that there is sufficient oxygen in the blood. Ions bring salt into the sea, control your thirst and
ionic constituents are used as food by all living things – from bacteria up to human beings.
A knowledge about the type and number of ions to be found in the environment helps us to
understand biochemical and ecological relationships. If the ionic concentrations in a foodstuff
is known then these provide us with information about whether the food is safe to eat or not.
There are many different ways of determining ions qualitatively (by their type) and
quantitatively (by their amount). Each piece of information is important. One method used for
obtaining this information is ion chromatography. Chromatography basically means "writing
with color". In traditional analysis this means the separation of substances according to their
color and their determination by visual observation. Al-though not all ions are characterized by
visible colors the term has been retained, but other methods of determination are used today.
Ion chromatography is one member of this large family of chromatographic methods. It can be used – to
put it very simply – to determine all ions which carry one or two charges. In the past ion chromatography or
"IC" used to be a very expensive method but today it is much more favorably priced. This is why it has
developed into a universal and powerful analytical tool which is easy to use.
This "Practical Ion Chromatography" will show that IC is not just an abstract analytical form,
but that it can provide rapid answers to common daily problems such as: Is the drinking water
suitable for feeding babies? How much nitrate is there in the spinach? Why does the washing
machine scale up? Does the wastewater cause environmental pollution? As accurate practical
analytical work is almost impossible without a theoretical background this monograph also
contains detailed information in a separate theoreti-cal section.
"Practical Ion Chromatography" is intended not only to provide you with a knowledge of
the basic prin-ciples of IC but also to provide you with an overview of general
chromatographic principles. And chroma-tography can do an awful lot: it satisfies
scientific curiosity and ensures healthy survival in a polluted environment.
6 Metrohm Monograph
3 Theoretical section
3.1 The history and importance of ion chromatography
The beginnings of ion chromatography (IC) or, more exactly, ion exchange chromatography go back to the
middle of the previous century. Between 1935 and 1950 knowledge about ion exchangers and their appli-
cations was considerably extended by the "Manhattan project". In the fifties and sixties theoretical models
for understanding the phenomenon of ion exchange and of ion chromatography, which is based on this,
were worked out. Continuous detectors were used in the seventies; this allowed the jump from low-
pressure to high-performance chromatography to be completed.
Table 1 History of ion exchange and ion chromatography, the analytical technique based on ion exchange
ca. 1850 Soil as an ion exchanger for Mg2+, Ca2+ and NH4+ Thomson & Way LC
1935 Sulfonated and aminated condensation polymers Adams, Holmes
(phenol/formaldehyde)
1942 Sulfonated PS/DVB resin as cation exchanger d'Alelio
(Manhattan Project)
1947 Aminated PS/DVB resin as anion exchanger McBurney
1953 Ion exclusion chromatography Wheaton, Baumann
1957 Macroporous ion exchangers Corte, Meyer, Kunin et al.
1959 Basic theoretical principles Helfferich
1967-70 Pellicular ion exchangers Horvath, Kirkland
1975 Ion exchange chromatography with conductivity Small, Stevens, Baumann
detection using a "stripper" HPLC
1979 Conductivity detection without a "stripper" Gjerde, Fritz, Schmuckler
1976-80 Ion pair chromatography Waters, Bidlingmeier,
Horvath et al.
The term 'ion chromatography' was coined in 1975 with the introduction of detection by conductivity
combined with a chemical reduction in conductivity by Small, Stevens and Baumann; it was subsequently
used as a trade name for marketing purposes for a long time. In the meantime the abbreviated term 'ion
chromatography' has become established as the superordinate term for the ion exchange, ion exclusion
and ion pair chromatography methods included under high performance liquid chromatography (HPLC)
[1]. IC today is dominant in the determination of anions while the atomic spectrometry
methods, com-monly used for the determination of cations, are hardly useful for determining
the electronegative anion formers of the fifth to seventh main groups of the periodic system.
The most important field of application today for anion chromatography is the routine investigation of
aqueous systems; this is of vital importance in the analysis of drinking water [2,3,4]. IC is also used for the
analysis of the element species in anionic elements or complexes; this is mainly for solving environmen-
8 Metrohm
Monograph
Figure 1 Division of chromatography
methods according to the
states of the stationary and
mobile phases
Retention parameters
If a mixture of substances is subjected to chromatographic separation then a distribution equilibrium
is formed between the mobile and stationary phases for each individual component. The
substances can only be separated successfully when the distribution coefficients D of the
components differ sufficiently from one another. D is defined as the ratio of the concentrations of a
substance A between the stationary (Index S) and the mobile phases (Index M):
(1)
Accordingly substances with a larger distribution coefficient D will be held back (retained) more strongly
than those with a smaller D. The chromatographic separation procedure is shown in the form of a chro-
matogram in which a detector signal is recorded as a function of the elution volume of the mobile phase or
the time. This means that it corresponds to a concentration or mass profile as a function of time. The
detector signal should be proportional to the concentration of an analyte at the end of the migration path
[8]. As shown in equation 2, the residence time or gross retention time tR of a substance on the stationary
phase is obtained by adding the net retention time tS, which corresponds to the actual residence time on the
migration path, and the flow time of the mobile phase without any interaction, the dead time tM.
(2)
Owing to the formation of channels, diffusion processes or irregularities in the equilibrium achieved be-
tween the mobile and stationary phases some particles may pass through the stationary phase more
slowly or more quickly than is to be expected from the net retention time t S. This means that a chromato-
gram does not consist of an infinite number of narrow signals, but ideally of Gaussian peaks (Figure 2).
10 Metrohm
Monograph
The width at half the peak height is known as the half width b 0,5 and corresponds to the 2.354-fold variance
σ of the distribution. The base width w is defined by the difference of the points of intersection
of the slope tangents with the y-axis which is the same as the 4-fold variance of the Gauss
function. Both quantities are a measure of the performance of a chromatography separating
column and, with an ideal peak shape, can be used for calculating the number of plates.
Variations from the ideal peak form can be described by a so-called asymmetry factor T. This
is defined as the ratio of the distances A and B between the central verticals and the slopes of
the distribution at 10% of their height (Figures 2 and 3) and can be calculated to:
(3)
For Gaussian peaks T = 1. A variation towards a larger T value is known as tailing, towards a smaller one
as fronting. In practice the intention is to achieve an asymmetry factor from T = 0.9 to 1.1.
(4)
At small values of k' a substance elutes close to the dead time or at the dead volume of the
chromatogra-phy system; this means that separation is poor. If k' is very large this means that,
although the separation is good, there is a long residence time on the migration path and the
peak becomes wider. Ideally the retention factor should be between 2 and 5.
Two substances will only be adequately separated if their retention factors differ from each
other suffi-ciently. The selectivity coefficient α, also known as the relative separating factor,
is a measure of the separability of two substances and is defined as follows:
(5)
If two substances cannot be separated then α = 1 and coelution occurs. The larger the value
of α, the better the separation. However, as α increases the time required for the separation
also increases, so that in practice selectivity coefficients of α = 1.5 are aimed for [10].
The selectivity coefficient does not describe the quality of the separation process. The resolution R does
not just take the relative positions of the peaks into account, but also their half-width (b 0.5) and base width
(w) values, as can be seen in Equation 6.
If the difference between the retention times of two peaks is large in relation to their base
widths or half-widths then the resolution is good. If an ideal peak symmetry is assumed then
two substances with R = 0.5 can still be identified. For qualitative separation R should be 1
(4σ-separation), for quantification a resolution of R = 1.2 to 1.5 is aimed for [25]. Resolutions
of R ≥ 2 (8σ-separation) are to be avoided because of the long analysis times involved.
(7)
Instead of the number of theoretical plates the height equivalent to a theoretical plate
HETP can also be used to describe the separating performance.
(8)
From Equations 5 ... 8 it can be seen that a stationary phase with a very large number of
theoretical plates can even separate substances from one another whose selectivity
coefficients or resolutions hardly differ at all. The equations also allow the calculation of
the number of theoretical plates which are necessary to solve a separation problem.
The theoretical separation stages model can be used to explain the occurrence of Gaussian signals
in chromatography if it is assumed that, because of flow and diffusion processes, only a finitely
quick and incomplete equilibrium is achieved between the mobile and stationary phases. This
results in a peak broadening process as a narrow substance zone at the start of the migration path
clearly becomes broader as the residence time on the stationary phase increases.
The calculation of the number of theoretical plates according to Equation 7 assumes that the
peak shape is ideal; however, this rarely occurs in reality. With asymmetric peak shapes the
calculation must be carried out according to the momental method [13]. Equation 9 includes
the symmetry factor and produces approximate values which make sense.
12 Metrohm
Monograph
(9)
A number of effective plates n, which represents the actual separation performance more closely
than the number of theoretical plates N, is corrected by the retention factor k' and is obtained from:
(10)
(11)
The three terms A, B and C depend in different ways upon the flow rate u of the mobile phase. The terms
A and B describe the whole mass transport through the stationary phase; term C is determined by interfer-
ence to the achievement of the equilibrium between the mobile and stationary phases.
Term A describes the eddy diffusion, which can be regarded as being a cause of peak widening owing to a
multi-pathway effect. This term is also known as the packing factor and is independent of the linear flow
rate u of the mobile phase, at least in a first approximation. The following relationship applies to term A:
(12)
(14)
Graphical representation of the van Deemter equation shows a hyperbolic curve, from
whose minimum value the flow rate u for the minimum plate height (maximum number of
plates) can be determined (Figure 4).
Even the dynamic theory is based on ideal requirements. In reality the three terms A, B and C
are only independent of one another in a first approximation, with there being an additional
influence of the flow rate u on the eddy diffusion (Term A). Term C can be differentiated by
using the terms CM and CS, which describe the mass transfer in the mobile phase (C M) and to
the stationary phase and back (C S). This is why the original van Deemter equation has been
modified for numerous applications in HPLC, GC and TLC [17,18].
14 Metrohm
Monograph
Modern liquid chromatography (LC)
Liquid chromatography LC is to be regarded as the generic term for numerous modern liquid
chromatog-raphy separation methods. It can be used for a wide range of different substances and is
characterized by its excellent analytical performance. LC also includes ion chromatography (IC)
which is probably the most important separation method used in modern analytical chemistry [3].
HPLC is a logical further development of the classical liquid chromatography (LC). In classical LC,
intro-duced by Tswett in 1906, glass columns with a diameter of 1 to 5 cm and a length of up to 500
cm were used; these were filled with separation phases with particle sizes of 150 to 200 µm. Even
separations of simple mixtures of substances could often take several hours with an average
separation performance. As a result of the understanding of the chromatography process which was
later developed (Equation 11) it became clear that increased performance could only be achieved
by a dramatic reduction of the particle diameter in the stationary phase; however, this placed
completely new demands on the equipment used for chromatography.
Since about 1970 special and powerful instrument technology has become available which is able to
overcome the high counter-pressures of 10 to 50 MPa which occur when packing materials with a particle
diameter of 3 to 10 µm and separating columns of 125 to 250 mm length x 4 mm ID are used.
As a result of the dramatic miniaturization HPLC has developed into a purely analytical
separating method; in contrast classical LC is today practically only used for preparative
purposes. The advantages of HPLC in comparison to classical LC are chiefly:
• excellent chromatographic efficiency
• continuous working process
• online detection of the separated substances
• high sensitivity and reproducibility
• utilization of the retention time for qualitative identification of substances
• short analysis times
Irrespective of its field of application, an HPLC system consists chiefly of the components shown in Figure
5: the high-performance pump with mobile phase (eluent) storage, injector (sample introduction),
sepa-rating column and detection system (including derivatization, data acquisition and processing):
Separating principles in LC
HPLC can be differentiated according to the different physicochemical interactions
between the substances in a sample and the stationary phase. Although in reality there
are usually several different mechanisms responsible for a successful separation [9], a
rough classification according to the following separation mechanisms is possible:
• adsorption
• distribution
• size exclusion
• affinity
• ion exchange
• ion pair formation
• ion exclusion
Adsorption chromatography is defined by interfacial reactions, in which liquid or gaseous
substances are enriched at a solid phase. Various models are available for providing a qualitative
and quantitative descrip-tion of adsorption processes; here we only refer you to the relevant
physical chemistry literature [19]. Two different techniques are used. In normal phase
chromatography the stationary phase is usually silica gel and therefore considerably more polar
than the mobile phase (hydrocarbons). In reversed phase chroma-tography the conditions are
exactly the opposite. For practical reasons, which chiefly concern eluent handling, virtually only RPC
is used today [3,9].
In distribution chromatography the stationary phase is a liquid which is immiscible with the mobile
phase. Separation is based on the differing solubilities of the analytes in the two phases. In an ideal
case the Nernst distribution law applies. This separating mechanism plays an important role,
particularly in gas chromatography when capillaries coated with separating liquids are used as the
stationary phase. Distri-bution chromatography may also occur in HPLC if silica gels modified with
non-polar hydrocarbons, e.g. so-called octadecyl phases, are used as the separating material.
Size exclusion chromatography (SEC) allows separation according to the molecular size as a result of
sieve effects. Silica gels or organic polymer resins with a defined pore structure are used as the stationary
phase. Smaller analytes can diffuse into the pores and are retarded. As the molecule size increases any
interaction with the pores becomes less likely, until at a particular size molecules are completely excluded
and practically elute in the dead volume. SEC is widely used in polymer analysis and bioanalysis.
Affinity chromatography allows the separation of mixtures of substances by selective or specific interac-
tive forces. Highly specific interactions can be observed between antibodies and antigens (key-keyhole
principle), as well as with enzymes and their substrates in particular. In practice enzymes or antibodies are
chemically immobilized on a stationary phase. If there is a corresponding substrate or antigen in the
16 Metrohm
Monograph
sample then this is retarded with extreme selectivity. This is why bioaffinity chromatography is
indispens-able in the active substance analysis sector (pharmacology).
Ion exchange chromatography (IC) together with ion pair and ion exclusion
chromatography are de-scribed in detail in the following section.
(16)
The different affinities of A – and B– to the functional groups mean that separation is possible. The equilib-
rium constant K is also known as the selectivity coefficient and is calculated as follows for anion A-:
(17)
If it can be assumed that the concentration of the eluent ions is normally higher than that of the
analyte ions by several powers of ten then [E–] can be regarded as being a constant in the mobile
and stationary phases. This means that the distribution coefficient D A (Equation 1) and the retention
factor k'A (Equation 4) can be calculated. Strictly speaking, such calculations are only permissible if
the concentrations in Equa-tion 17 correspond to the activities; however, this is only the case for an
infinite dilution [19]. In principle the activities of the ions in the stationary phase are inaccessible [4].
For the most frequently used ion exchangers of low capacity, which can only be used as the mobile
phase with very dilute electrolytes, the activities are simply disregarded. Such very coarse
approximations are no longer valid for high-capacity (> 200 mmol/g) packing materials and
concentrated eluents; these show clear variations from the 'ideal' behavior.
18 Metrohm
Monograph
3.3.3 Ion pair formation
With the aid of ion pair chromatography it is possible to separate the same analytes as in ion
exclusion chromatography, but the separation mechanism is completely different. The stationary
phases used are completely polar reversed phase materials such as are used in distribution
chromatography. A so-called ion pair regent is added to the eluents; this consists of anionic or
cationic surfactants such as tetraalkylammonium salts or n-alkylsulfonic acids. Together with the
oppositely charged analyte ions the ion pair reagents form an uncharged ion pair, which can be
retarded at the stationary phase by hydropho-bic interactions. Separation is possible because of the
formation constants of the ion pairs and their different degrees of adsorption. Figure 7 shows a
simplified static ion exchange model in which it is assumed that interactions with the analytes only
occur after adsorption of the ion pair reagent at the stationary phase.
Figure 7 Schematic diagram showing the static ion exchange model in ion pair
chromatography (IPC). The separation principle applies to both anions and cations.
In IEC a completely sulfonated cation exchanger whose sulfonic acids groups are electrically neutral with protons
as counter ions is frequently used as packing material. In aqueous eluents the functional groups are hydrated. The
hydrate shell is limited by an (imaginary) negatively charged membrane (Donnan mem-brane). It is only passable
by uncharged, non-dissociated molecules such as water. Organic carboxylic acids can be separated if strong
mineral acids such as sulfuric acid are used as the mobile phase. Owing to the low acid constants (pK A values) of
the carboxylic acids these are present in a virtually fully non-dissociated form in strongly acidic eluents. They can
pass through the Donnan membrane and be adsorbed at the stationary phase, whereas the sulfate ions of the
completely dissociated sulfuric acid are excluded.
Figure 9 shows the typical dependency of the elution volume of an acid on its pKA value
for separation by ion exclusion. Superimposed adsorption (long-chain carboxylic acids,
H2S) and the limits of the practical working range can be clearly recognized. In the final
instance carboxylic acids are separated because of their different pKA values.
20 Metrohm
Monograph
3.4 Retention models in ion chromatography
In an ideal case the retention of an analyte in ion chromatography would only be determined by its affinity
to the functional groups of the ion exchanger. This affinity can be described by the formulation of a chemi-
cal reaction, the ion exchange reaction, and can be explained by using the law of mass action.
The retention models described below attempt to make predictions about the retention behavior of partici-
pating analytes under particular chromatographic conditions based on the law of mass action. If the
resulting models are suitable for explaining the macroscopic observations then with their help it is pos-
sible, for example, to optimize an elution system for a particular separation problem.
(18)
(19)
As the activities of the participating ions cannot be determined in the stationary and
mobile phases the activity in the stationary phase is ignored and set as being 1.
If for the analyte anion Ax– two known quantities, the distribution coefficient DA and the
retention factor k'A, are now introduced from section 3.2.1,
(20)
(21)
As the concentration of the eluent ions E is normally higher than that of the analyte
anions Ax– by several powers of ten, a good approximation can be obtained by assuming
that all the functional groups are occupied by Ey–. Under this assumption the non-
determinable concentration of Ey– in the stationary phase can be replaced by the more
easily accessible parameters of exchange capacity Q and charge of the eluent anion y:
(22)
The retention factor k'A of the analyte anion Ax– can easily be obtained from a
chromatogram. Equation 23 is therefore solved for this quantity.
(24)
(25)
22 Metrohm
Monograph
As a first approximation it can be assumed that the selectivity coefficients are
independent of Q at constant Φ; this results in the following proportionality:
(26)
From Equation 26 it can be seen that if the exchange capacity Q is increased then the
concentration of the eluent [E y–] must be increased proportionally in order to obtain constant
retention factors. This is the reason why low-capacity separation phases are normally used in
ion chromatography, as high electrolyte concentrations would make the most important
detection method used in ion chromatography, conductiv-ity detection, practically impossible.
In order to optimize separation problems the eluent concentration [E y–] is often varied. If all
the other parameters occurring in Equation 25 are kept constant then this can be simplified to:
(27)
A graphical plot of Equation 27 gives a straight line with a slope m = – x/y and an
intercept on the axis C which contains the quantities Q, Φ and K A,E. If a monoanionic
eluent is used then m is also known as the effective charge. Figure 10 shows the result
of Equation 27 for various combinations of differently charged eluent and analyte anions.
The graphical representation of this equation is similar to Figure 10, but with a positive slope. Chromato-
graphic investigations into the variation of Q have to date only been carried out once for the separation of
divalent cations. This has shown that, in contrast to previous assumptions, the retention factor and the
selectivity coefficients cannot be regarded as being independent of the exchange capacity. For the optimi-
The above considerations only apply for one analyte anion. If two different anions A x– and Bz–
compete for the functional groups then the following applies for the selectivity coefficients K A,B:
(29)
(30)
(31a)
(31b)
which can be simplified to for analytes with the same charge (x = z):
(32) or (33)
For the selectivity between two similarly charged analyte anions this means:
• it is only a function of the selectivity coefficients KA,B and the charges z and x,
• at constant KA,B the selectivity depends neither on the concentration [E y–] nor on the
chemical consti-tution of the eluent anion (!)
If A and B have different charges then:
• a depends on the retention factor of one of the two analytes,
• the two retention factors k'A and k'B are not independent of one another (!)
In Equations 31 to 33 it is particularly interesting that the selectivities of two anions initially depend on
neither the chemical constitution nor on the charge of the eluent anion, providing that the phase-volume
ratio and the selectivity coefficient are constant. However, in practice an alteration in a can be achieved by
a variation of [E y–], as two analytes with the same charge can nevertheless have different chemical
properties, e.g. polarizability and hydration; this can result in different affinities to the stationary phase.
However, these interactions are not taken into account in the classical derivation.
24 Metrohm
Monograph
Retention models for eluents with several anions
The previous observations have referred to elution systems with only a single eluent
anion. In practice there are usually several eluting species present, for example in
carbonate/hydrogen carbonate buffers or in multi-base acids such as phosphoric acids,
whose dissociation and therefore distribution of species depends strongly on the pH.
Even in simple cases in which none of the participating eluent anions is involved in the acid-base
equilib-rium, the relationship between the retention factor k' and the eluent concentration [E –] cannot
be repre-sented in the form of a simple log-log relationship according to Equation 28. This would
only be possible if the concentration or the elution power of the other eluent anions could be
ignored; this would then correspond to the retention model for monoanionic eluents.
In the literature several models concerned with polyanionic eluents are described; these
are briefly dis-cussed below:
• dominant equilibrium model [21]
• effective charge model [22-24]
• multiple eluent species model [25, 26]
If an eluent based on phosphate with H2PO4–, HPO42– and PO43–, (otherwise H2P–, HP2– and
P3–) and the monovalent analyte ion A– is considered then the following equilibria are formed:
(34)
(35)
(36)
Here the quantities x1..3 correspond to the shares of the particular reaction in the retention, which is why:
(37)
Both the dominant equilibrium model and the effective charge model postulate a particular
charge for the eluent anion, even though several species are present; this means that the
retention model for monoanionic eluents obtained in Section 3.4.1 can be used.
The dominant equilibrium model assumes that the equilibrium in Equation 36 is wholly on the right-hand
side, as P3– is bound far more strongly to the stationary phase than H 2P– and HP2– as a result of its higher
charge. This means that P3– alone is decisive for the elution so that the charge of the eluent anion is –3.
However, in practice this model only achieves a good agreement with multivalent analytes [4].
In the effective charge model an effective charge is calculated, taking into account the pH value, from
the molar fractions of the possible species H 2P–, HP2– and P3– [22]. By using these together with the
existing concentrations of the eluent species a relationship analogous to Equation 27 can be obtained.
However, a requirement for such a type of calculation is that the selectivities of the eluent species do not
differ greatly
(38)
(39)
The further mathematical treatment is carried out as for the derivation of the retention
model for monoanionic eluents. The following must chiefly be taken into account:
• the (possible) dissociation of the analyte anion A–
• the total concentration of the eluent species: cP = [H3P] + [H2P–] + [HP2–] + [P3–]
• the extent of the interactions between the eluent species and the functional groups
The introduction of the retention factors k' A (Equation 20) and the capacity Q (Equation
22) supplies, after further mathematical conversion, a complicated expression for k'A [28];
this is given here only in its logarithmic and further simplified form:
(40)
C3 is a constant which, similar to Equation 27, contains quantities such as the phase-volume ratio,
the capacity and the equilibrium constant; c P is the total of the concentrations of the eluent species.
From Equation 40 it can be deduced that the slopes of the straight lines in a double logarithmic plot
must always be smaller than those according to the simple retention model for monoanionic eluents
(Equation 27), as the total in the brackets is always smaller than one. It is also clear that the pH
value has a decisive influence on the extent to which the log-log relationship is influenced.
For eluent species which are not chemically derived from one another Janoš et al. have provided a model
which was developed for describing eluents containing a phosphate buffer and additional perchlorate [29].
This model was derived according to similar considerations to those described above, but in addition an
exchange equilibrium must be taken into account for a further monovalent eluent ion. The calculations
provide very complicated expressions for the retention factor; these can be dramatically simplified for
neutral or acidic eluents. If only a single further monovalent eluent species is present in addition to the
perchlorate then Equation 41 is obtained in which x and y represent the contributions of the corresponding
equilibrium reactions (x: phosphate buffer, y: perchlorate) to the retention. As in the other models, C is a
26 Metrohm
Monograph
constant, while the factor a, which is not defined any more closely, is intended to take
into account how much more strongly the perchlorate ion is bound at the stationary
phase than the phosphate species involved.
(41)
As in Equation 41 the terms within the brackets are always smaller than one, the slope
of the log-log plot is always less than would be expected from the simple retention
model. In actual applications the model provides good agreement with the experimental
data. However, the form described above cannot be used for alkaline elution systems.
(43)
(44)
In order to take the influence of the complexing agent on the ion chromatography separation into
account the retention model for isoionic displacement (see Section 3.4) is extended. The α M value is
introduced as an influencing quantity which describes the degree of complex formation of the
analyte. The fraction αM of the free analyte ions in the mobile phase is given as
(45)
with [Me'] as the total concentration of the metal ions. The α M value can be calculated from the complex
formation constants, the acid dissociation constant of the carboxylic acid and the pH of the eluent. If
complex formation is taken into account then the following is obtained for the distribution coefficient D Me:
(46)
If it is assumed that only free analyte ions Mex+ interact with the carboxylic acid or sulfonic acid
groups and that c(Ez+) >> c(H+), then the following is obtained for Equation 21:
(47)
(48)
If several cationic metal species occur together, e.g. Me x+ and MeHL(x–1)+, then normally only a single
peak is obtained in the chromatogram for the analytes involved. The number of peaks which are
obtained depends on the kinetics of the complexing and decomplexing equilibria in the mobile
phase. Only one peak is obtained if the complex equilibria are achieved more rapidly in the mobile
phase in comparison to the residence time of the complex in the stationary phase. On the other
hand, if the complexing process occurs slowly then asymmetric or multiple peaks may occur.
28 Metrohm
Monograph
If it is assumed that all metal species present in the mobile phase can interact with the stationary phase
then the following is obtained for the experimentally determined capacity factor k' exp of the analyte:
(49)
Consideration of the dependency of the capacity factor on the influencing quantities Q, [E y+]
as well as αM requires that the relationship presented in Equation 48 is used as the basis, as
the divalent analytes mainly form neutral or anionic complexes with strong complexing agents.
Calculation of αM values
According to Equation 45, the αM value is defined as the ratio of the concentration of the
free metal ions to the total concentration of the metal ions. The concentrations of the
metal species present in the mobile phase can be calculated from the relevant complex
formation constants and the acid dissociation con-stants of the carboxylic acids used.
If tartaric acid is used as the complexing agent in eluents then it is chiefly the neutral
MeL 1:1 complexes which are formed with the alkaline earth, transition and heavy metals
together with a lesser amount of the hydrogen tartrate complex MeHL +. For tartaric acid
eluents the following is obtained for the calculation of the αM value:
(50)
where cL is the total concentration of tartaric acid and αHL and αL are the molar fractions
of the acid anions HL– and L2–.
Apart from 1:1 complexes, some metal ions also form stable MeL 22– complexes with
oxalic acid and/or pyridinedicarboxylic acid , so that αM can be calculated as follows:
(51)
(52)
(53)
αM value are obtained from the laws of mass action of the individual deprotonization stages:
(54)
(55)
(56)
Conductivity detection
Conductivity detection, also known as conductometric detection, has a market share of about
55% in the ion chromatography sector [4]. If the number of ion chromatographs which have been
sold is taken into account then this share is probably much higher today. Conductivity detection is a
non-selective detection principle; in this case both direct and indirect detection determinations are
possible. As aqueous electro-lytes are frequently used as the mobile phase in ion chromatography,
the detector must be able to respond to the relatively small changes in the total conductivity of the
eluent caused by the analyte ions. By the use of so-called suppression techniques the self-
conductivity of particular eluents can be dramatically re-duced; in the case of strong acid anions it is
also possible to achieve a considerable improvement to the sensitivity.
30 Metrohm
Monograph
The conductivity κ is determined technically as the reciprocal of the resistance R which a
liquid produces between two electrodes with an area of A at a distance of L.
(57)
The limiting conductivity Λ∞ and the variation of conductivity with concentration can be
determined by using Equation 59. Constants A and B are empirical substance constants.
(59)
The conductivity of an electrolyte is obtained by adding the ionic conductivities Λ –Anion and Λ+Cation together:
(60)
(61)
(62)
In ion chromatography the conductivity of an eluate can either be determined directly or after passage
through a suppressor. These versions are known as the single column and suppressor techniques. The
version which is to be preferred can be determined by carrying out a rough calculation.
If direct conductivity detection is used in anion chromatography then the sensitivity κ Peak of the
measure-ment depends on the difference in the equivalent conductivities of the
analyte and eluent anions; with chloride as the analyte and carbonate as the eluent
anion the following equations are obtained:
κ Peak ≈ c Analyte
(Λ-– – – Λ
-– 2– ) ⇒ κ Peak
≈c Analyte
(76 – 72)
Cl
CO3
κ ≈c •4
Peak Analyte
If the eluent is adapted to the requirements of direct conductivity detection then the
following sensitivity is obtained by replacing the carbonate eluent by a phthalate eluent:
κ Peak ≈ c Analyte
(Λ-– – – Λ-– ) ⇒ κ Peak
≈ c Analyte (76 – 38)
Cl Phthalate
κ ≈c • 38
Peak Analyte
If, on the other hand, the conductivity of the eluent is chemically suppressed (exchange of the
eluent cations for H+), the sensitivity depends on the sum of the equivalent conductivities of
the analyte anion and the H+ ion; the following then applies for Cl– as analyte anion:
κ Peak ≈ c Analyte + +)
(Λ-– – + Λ ⇒ κ Peak ≈ c Analyte (76 + 350)
Cl H
κ ≈c • 426
Peak Analyte
From this rough calculation it can be seen that, for anions, direct conductivity detection is
less sensitive by a factor of 10 than conductivity detection after chemical suppression.
32 Metrohm
Monograph
In a similar way the following rough calculation can be made for cation chromatography, with
Na+ as the analyte and H+ as the eluent cation. In the case of direct conductivity detection
(NaCl/HCl) the following equations are obtained for the sensitivity κ Peak of the measurement:
+ +
κ Peak
≈cAnalyte
(Λ + –Λ + )⇒ κ
Peak Analyte
≈c (50 – 350)
Na H
κ ≈c • (–300)
Peak Analyte
If, on the other hand, the conductivity of the eluent is chemically suppressed (exchange
of the eluent anions Cl– for OH–), the following applies:
κPeak ≈ cAnalyt (Λ+Na+ + Λ-–OH–) ⇒ κPeak ≈ cAnalyt (50 + 198)
κ ≈c • 248
Peak Analyt
This means that the sensitivity is better for direct conductivity detection of cations than
for conductivity detection after chemical suppression.
Table 2 Equivalent conductivity Λ∞ of several
ions
The membrane suppressor shown in Figure 15 allows continuous work but, as a result of the
use of ion exchange membranes, is susceptible to occupation of the membrane surface; this
reduces the suppres-sion capacity and finally causes the suppressor to cease functioning.
Amperometric detection
In principle voltammetric detectors can be used for all compounds which have functional groups
which are easily reduced or oxidized. The amperometric detector is the most important version. In
this detector a certain potential is applied between a working electrode and a reference electrode. If
an electrochemi-cally active analyte, whose half-wave potential is such that the applied potential
causes reduction or oxidation, now passes between the electrodes then a current will flow; this
represents the measuring signal. Amperometry is very sensitive; the conversion rate is only about
10%. Apart from a few cations (Fe 3+, Co2+) it is chiefly anions such as nitrite, nitrate, thiosulfate as
well as halogens and pseudohalogens which can be determined in the ion analysis sector. The most
important applications lie, however, in the analysis of sugars by anion chromatography and in
clinical analysis. Owing to its different working prin-ciple, the coulometric detector provides a
quantitative turnover without, however, there being any in-creased sensitivity.
34 Metrohm
Monograph
Potentiometric detection
In potentiometric detection ion-sensitive electrodes are used, some of which have a very high
selectivity. Despite their necessary high degree of miniaturization the sensors must function reliably;
this still causes problems in practice. This is the reason why up to now potentiometric detection, in
the ion chromatogra-phy sector, is limited to a few special applications.
Photometric detection
Because of its extremely wide range of application photometric or UV/VIS detection is the most
important detection method used in HPLC, as virtually all organic molecules contain chromophore groups
which are able to absorb in the UV or VIS spectrum. A requirement is that the eluent used does not absorb
in the range of wavelengths used. With direct detection at the maximum absorption of an analyte UV/VIS
detec-tion is practically selective. Substances which either only have a limited absorption or no absorption
at all in the particular wavelength range can be determined indirectly by measuring the maximum
absorption of the elution system. In the field of inorganic ion analysis UV/VIS detection plays a smaller
role. While of the simple anions only analytes such as nitrate, bromide or iodide absorb, important
analytes such as fluoride, sulfate or phosphate can only be measured indirectly [4]. Many cations do not
absorb at all, but multivalent and transition metals in particular can be converted in a post-column
derivatization with chelate formers such as 4-(2-pyridylazo)-resorcinol (PAR) or Tiron to form colored
complexes. Redox-active analytes such as bromate and other oxohalide ions can be analyzed by UV/VIS
detection after undergoing a post-column reaction with an electrochemically active indicator.
Fluorescence detection
Fluorescence detection is very sensitive and is always possible whenever analytes can be excited
to fluoresce; this is mainly the case for organic compounds with extended π-electron systems. This
means that typical applications are found in the fields of clinical and organic analysis. In connection
with ion chromatography, fluorescence detection is used in a few special cases, as only particular
ions such as Ce3+ are directly accessible and non-fluorescing ions can only be detected after
derivatization. It is ex-tremely difficult to develop elution systems for this detection method because
of its great susceptibility to interference by contaminants. Furthermore the linear range of the
method is relatively small (often less than two powers of ten) owing to self-absorption effects.
Coupling techniques
So-called coupling techniques represent the link-up of a chromatography system with an
independent analytical method, usually spectrometry [3]. In recent years these methods have
greatly increased in importance. Although the coupling of a gas chromatograph with a mass
spectrometer (GC-MS) is well established, coupling HPLC with spectrometric methods causes great
technical problems. In classical HPLC, i.e. the analysis of organic compounds, couplings with a
mass spectrometer (LC-MS), IR-spec-trometer (LC-FTIR) and nuclear magnetic resonance
spectrometer (LC-NMR) are available [3]. In particu-lar, powerful atomic spectrometric detectors are
used in ion chromatography (IC). Examples are atomic emission and mass spectrometry with
inductively coupled plasma (IC-ICP-AES, MS); as a result of their element specificity and sensitivity
these provide excellent performance data. This is the reason why, de-spite their relatively high
costs, such systems are used for the analysis of species and in the ultratrace analysis of elements.
36 Metrohm
Monograph
All stationary phases can be further differentiated according to their type of application
(anion or cation chromatography) or the structure of the functional group. The packing
materials based on silica gel were originally used for ion chromatography. Although they
have a very good separating performance and are mechanically extremely stable, their
chemical instability means that they can only be used in the pH range 2 to 7.
From about 1980 onwards ion exchangers became available which were based on organic
polymers and could be used in ion chromatography; these were manufactured by modifying
commercially available adsorber resins. Today the packing materials used are normally based
on either polystyrene-divinylbenzene copolymers (PS-DVB) or methacrylate polymers (MMA).
These two basic copolymer types mainly differ in their polarity. The PS-DVB copolymers
are completely non-polar and represent RP phases whereas the MMA polymers are
relatively polar. This situation is an advantage in IC as the more polar separation phases
have a lesser tendency to secondary interactions such as adsorption.
The greatest advantage of the organic polymer resins is their great chemical stability throughout the
whole pH range. After initial problems had been overcome their chromatographic efficiency is
similar to that of the silica gels. However, MMA phases may have a limited mechanical stability, this
could limit the length of the separation column used or the maximum possible eluent flow rate.
In ion chromatography today two types of stationary phases are used which differ in principle;
surface functionalized ion exchangers and pellicular ion exchangers. In the first type the functional
groups are located directly on the polymer surface or in the pores; the pellicular materials have very
small particles (also surface functionalized) which are bound to larger central particles [4]. The
bonding can either be mechanical or result from hydrophobic or electrostatic interactions. Figure 17
shows the arrangement of the two types of packing material using anion exchangers as examples.
Figure 17 Structure of surface functionalized (a) and pellicular anion exchangers with mechanical bonding (b)
The pellicular packing materials have a higher chromatographic efficiency as the diffusion
paths are kept very short owing to the greater distance of the functional groups from the base
material; this results in excellent mass transfer. However, the chemical stability of these
separation phases is considerably less than that of the surface functionalized materials.
Figure 18 Overview of the most important functional groups which are mentioned in this work
TMA: Trimethylamine (Type I) DEMA: Diethanolmethylamine
EDMA: Ethyldimethylamine TEA: Triethanolamine
DMEA: Dimethylethanolamine (Type II)
In commercial materials, which are usually derived from Type I or II, the exact
arrangement of the func-tional groups is a well-guarded secret [4].
38 Metrohm
Monograph
3.6.4 Cation exchangers based on silica gel
With cation exchangers based on silica gel a differentiation is made between directly
functionalized mate-rials and those which are coated by a polymer.
Practically the only reports concerning functionalized materials are about strongly acidic exchangers with a
sulfonic acid group [a2,a4]. These have a good chromatographic efficiency, but are unsuitable for the
simultaneous determination of alkali metals and alkaline earths owing to the large differences in affinity.
With polymer-coated silica gels, the so-called Schomburg phases, the silicate surface is coated with
a "prepolymer" which is then immobilized by cross-linking. By subsequent functionalization various
different types of exchanger can be realized. In order to obtain weakly acidic cation exchangers
polybutadienemaleic acid (PBDMA) is used, this is then radically cross-linked in-situ [30]. As a result
of the thin polymer layer of about 1 to 5 nm [31] the analyte diffusion paths are short so that a high
degree of chromatographic efficiency is the result.
A large number of applications for silica-gel-based ion exchangers exist. The simultaneous
separation of alkali metals and alkaline earths is one of the most frequent applications; the
separation of transition metal and heavy metal ions is also possible. However, several
disadvantages prevent the universal use of ion exchangers based on silica gels:
• At pH < 2 the bond between the silicon matrix and the functional group becomes increasingly
weaker; this leads to a gradual erosion of the functional groups and a loss in capacity.
• At pH > 7 the solubility of the silica gel increases considerably; this results in a reduction in the
mechanical stability of the packing, particle break-up and a dead volume at the start of the column.
The above definition of the capacity is based on a complete equilibrium being attained between the
station-ary and mobile phases; this is why it is also known as the static capacity. In contrast, the
dynamic (effective) capacity is understood to be the number of functional groups which are actually
available during a chromatographic process. It is always smaller than the static capacity [2,4].
The exchange capacity can be determined in different ways [33]:
• volumetrically or titrimetrically
• by elemental analysis
• by determining retention times
All the methods provide different values for Q for the same material. Volumetric methods are most fre-
quently used in practice. For anion exchangers a packed separating column or a defined amount of resin
is loaded, e.g. with chloride solution. After the residual chloride has been washed out the column or resin
can be eluted with nitrate. The amount of chloride eluted, which corresponds to the static exchange
capacity under equilibrium conditions, can be titrated against AgNO 3 solution and quantified.
The pH dependency of ion exchangers can usually be neglected. With cation
exchangers the capacity of the weakly acidic carboxylic acid groups (R-COOH) is only
given at higher pH values (deprotonation), while the strongly acidic sulfonic acid groups
(R-SO3H) are always completely deprotonated and provide a pH-independent capacity.
40 Metrohm
Monograph
In combination with the retention models (Section 3.4) the importance of Q for the selection of the
elution and detection systems becomes obvious. Universal conductivity detection was practically
impossible to realize with eluents with a high ionic strength, no matter which special technique was
used. This is the reason why, since the introduction of ion chromatography in 1975, virtually only
low-capacity separating columns have been used. Up to now high-capacity ion exchangers have
only been described in IC appli-cations in combination with detection techniques which are not so
strongly limited by the elution system, such as UV/VIS detection [2,4].
While low-capacity separating materials are very suitable for the analysis of samples with a low
ionic strength and therefore a low matrix load, at higher ionic strengths they rapidly reach their
limits. Overload effects occur because of their low number of functional groups; these result in
deformed peaks and a dramatic drop in efficiency. Problems also occur if the analytes are present
in very different concentra-tions; this is almost always the case in ultratrace analysis.
Figure 19 Structures of the most important aromatic carboxylic acids used with the single column technique.
This class of substances is used so frequently as the solutions of the acids and their salts have a very high
elution power combined with a relatively small self-conductivity, so that they can be used for direct con-
ductivity detection. With multi-base acids the charge and therefore the elution power can be controlled via
the pH. However, this must be maintained very accurately. Aromatic carboxylic acids have a high absorp-
tion in the UV spectrum so that they can be used advantageously in indirect UV/VIS detection.
For aromatic sulfonic acids such as p-toluene sulfonic acid the same statements apply in
principle; these are always present in a deprotonated state and therefore possess no
buffer capacity. Their elution power can only be controlled via the concentration.
Aliphatic carboxylic acids such as oxalic acid and citric acid have a high self-conductivity but are UV-
transparent, so that they can be used for direct UV/VIS detection. This also applies to aliphatic sulfonic
acids, of which methanesulfonic acid is the one which is most frequently used. The higher homologues of
both classes of compounds with their long hydrocarbon chains and the low equivalent conductivities with
which this is associated offer the possibility of being used for direct conductivity detection [2,4].
Alkali hydroxides have only a very limited use in the single column technique, as the OH – ion has the
lowest affinity to quaternary ammonium groups of all possible eluent anions. As a result high eluent
concentra-tions are required even at low capacities so that at the best only indirect conductivity detection
can be used. In contrast, UV/VIS detection is easily possible, although the basically UV-transparent
hydroxide ion has a distinct absorption below 220 nm at higher concentrations.
42 Metrohm
Monograph
If inorganic acids or their salts are used then their virtually complete dissociation and associated high
conductivity means that only photometric detectors can be used. If phosphoric acid or phosphates are
used then the buffer capacity and the elution power can be controlled via the pH of the eluent.
Most of the eluents mentioned here also allow the use of other detection systems such as
amperometry, fluorimetry or coupling techniques. If you wish to use the elution systems
mentioned above with these detectors please refer to the relevant literature [2,4,5].
Suppressor technique
The suppressor technique is the detection technique originally used when IC was introduced [1]. In con-
trast to the single column technique, this method uses only conductivity detection. In the suppressor
technique a so-called suppressor is inserted between the separating column and the detector‚ which is
why this technique is also known as «suppressed ion chromatography» [2,4]. Both the eluent and the
analytes are chemically modified in the suppressor, this is of particular importance with respect to the
subsequent conductivity detection. The suppressor has the task of reducing the self-conductivity of the
eluent and, if possible, of increasing the detectability of the analytes.
The principle of chemical suppression is shown in Equations 63 and 64 for an anion
chromatography application. An NaHCO3 eluent is used, the chloride ion is the analyte.
Suppression is carried out with a strongly acidic H+-form cation exchanger.
(63)
(64)
In the simplest case the suppressor unit consists of a column inserted after the separating column,
which is where the older term «two-column technique» originates. The eluent sodium hydrogen
carbonate is neutralized according to Equation 63, as the sodium ions are replaced by protons. This
dramatically lowers the self-conductivity of the eluent. The analyte Cl – itself is not altered (Equation
64), but its counter-ion Na + is exchanged for H+, which has a considerably larger equivalent
conductivity [19]. As the detector records the sum of the conductivities of the analyte and counter-
ion as the signal, the two reactions which take place result in a clearly improved sensitivity.
The columns with H+-form cation exchangers originally used as suppressors made a
considerable contri-bution to broadening the peaks. In addition they could only be operated
discontinuously, as the cation exchanger had to be regenerated. This is why continuously
operating membrane suppressors are mainly used today; in these the regenerating agent,
usually dilute sulfuric acid, and eluent are led past each other in a counterflow principle [6].
Despite its advantages the suppressor technique also has several crucial disadvantages. In practice only
those eluents which are based on alkali hydroxides or carbonates can be successfully suppressed in
anion chromatography. Anions of weaker acids, e.g. acetate or fluoride, are almost completely protonated
after the suppression reaction, so that their detectability is much lower when compared with the single
column technique. High-valency cations must be removed before the analysis as they form insoluble
hydroxides which precipitate out on the separating column and can block it.
As has already been made clear, the suppressor technique is synonymous with chemical suppression of the
eluent and the use of direct conductivity detection [2,4]. This technique is unusually widespread in anion
chromatography, chiefly because in many cases it is more sensitive than the direct conductivity detec-tion used in
the single column technique. The self-conductivity of the classical eluents (e.g. 2 mmol/kg
• Alkali hydroxides
• Alkali carbonates and hydrogen carbonates
• Borates (e.g. B4O72–)
• Amino acids
In practice, of the eluents mentioned above only the alkali hydroxides and carbonate buffers are of
any great importance, which means that the choice of potential mobile phases is very limited.
The hydroxide ion is an extremely weak eluent ion, so that even with low-capacity separating
materials work must be carried out at high concentrations above 50 mmol/kg. If very polar
functional groups are used then the relative retention force of the OH – ion can be considerably
increased by utilizing the hydrox-ide selectivity. With OH – eluents manipulation of the retention
times or selectivities can only be carried out via the concentration.
The use of alkali carbonates and alkali hydrogen carbonates permits considerably more flexible elution
systems. Both species, HCO3– and CO32–, are present as carbonic acid H 2CO3 after suppression; this is
only dissociated to a very small extent. In its elution power hydrogen carbonate is only slightly stronger
than hydroxide, whereas carbonate is a relatively strong eluent. Both anions are normally used together;
this results in an eluent with a buffer effect whose elution power can be easily controlled via the concentra-
tions of the two components and the ratio of their concentrations. Because of the charges on the eluent
species the selectivity for monovalent and divalent analytes can be influenced very selectively. The con-
centration ratio of the two eluent ions can also be adjusted very accurately via the pH, this is why the pH
range used for HCO3–/CO32– eluents is between 8 and 11. Just as for OH – eluents, elution can be acceler-
ated by the use of stationary phases with polar functional groups.
Both elution systems can only be used successfully with surface functionalized anion
exchangers when either the matrix (methacrylate copolymers) or the functional group has a
high polarity. For separating columns based on PS-DVB very poor peak symmetries and long
retention times have been observed for weak analytes (nitrate, bromide) even when polar
functional groups are used. With pellicular materials these effects are not so marked, which
explains their exceptionally widespread use in the suppressor technique [2,4].
3.7.2.1 Cation chromatography of alkali, alkaline earth and ammonium ions with conductivity detection
In the separation of alkali metal and ammonium ions as well as short-chain aliphatic amines on sulfonated
separation phases by ion chromatography, mineral acids such as HCl or HNO 3 are mainly used as the
eluent [4]. The acid concentration of the eluent depends on the type and capacity of the cation exchanger
used and is a few mmol/L. Divalent cations such as the alkaline earths cannot be eluted with mineral acids
as they have a distinctly higher affinity to the stationary phase and a dramatic increase in the acid concen-
tration would make detection too insensitive or mean that suppression could no longer be guaranteed. As
an alternative an organic base such as ethylenediamine can also be used for the separation of alkaline
earth ions. At low pH values this is protonated and is present as a divalent cation.
The simultaneous analysis of alkali and alkaline earth cations on strongly acidic cation exchangers is
mainly carried out with eluents which contain hydrochloric acid and 2,3-diaminopropionic acid [2]. By
44 Metrohm
Monograph
varying the pH it is possible to alter the degree of protonation of the amino groups and
therefore the elution power of the 2,3-diaminopropionic acid (see section 3.4.2).
In ion chromatography systems without suppression, weak organic acids can also be
used as eluents in addition to the mineral acids. Organic acids used are e.g. oxalic acid,
citric acid, tartaric acid. Complexing agents such as 2,6-pyridinedicarboxylic acid
(PDCA) and 18-crown-6 are used to selectively influence the analysis times of individual
cations.
There are two different types of conductivity detection. If the basic conductivity of the eluents is high, e.g.
with dilute mineral acids because of the high conductivity of the H + ions, then direct conductivity detection
is possible with the analytes having a clearly lower conductivity than the eluent. This means that negative
peaks are produced which, however, can be converted into a normal chromatogram by reversing the
polarity of the detector or by simple inversion. If the quality of the conductivity detector does not permit
sensitive conductivity measurements at a high level then suppression is also used for cation chromatog-
raphy. The construction of suppressors for cation chromatography is far more difficult than for anion
chromatography and they also have a much shorter working life.
3.7.2.2 Cation chromatography of transition metal and alkaline earth ions with
post-column derivatization and photometric detection
Monovalent cations such as H+ or Na+ are unsuitable for use as an eluent for the separation of transition
and heavy metal ions as the selectivity coefficients of the analytes hardly differ at the same charge
number. However, separation can be realized by the introduction of a secondary equilibrium. This is done
by using complex-forming carboxylic acids (see Figure 20) such as citric acid, oxalic acid and tartaric acid
as the eluent. Together with the metal ions these form neutral or anionic complexes (see section 3.4).
By complexing the metal cations the effective charge density of the analytes is reduced. In addition
the different complex forming constants of the individual metal ions increase the selectivity of the
separation. The elution mechanism is a result of the isoionic displacement by the counter-ion
(pushing effect) and complex formation (pulling effect) by the complex-forming ligands [4].
Figure 21 shows the equilibria between the analytes M 2+, the complexing agent L2– and
the counter-ions E+ which participate in the elution mechanism.
The extent of the isoionic displacement by the counter-ions E + is determined by the affinity of the cation to
the stationary phase. With monovalent counter-ions the cation with the greater affinity to the stationary
phase has a stronger eluting effect. The influence of the complexing agent can be varied by altering the pH
and concentration of the eluent. In addition the elution power can be influenced by the use of several
complexing agents as well as by using a divalent cation as the counter-ion.
Whereas both types of conductivity detection are suitable for the detection of alkali metals and alkaline
earths, transition and heavy metal ions can only be determined by direct conductivity measurement with-
out suppressor. If the suppressor technique were to be used then these metals would be converted into
(mostly) insoluble hydroxides by the suppressor reaction. Direct conductivity detection is only possible
when eluents with a low basic conductivity are used together with low-capacity cation exchangers.
This is why post-column derivatization with the formation of photometrically detectable metal
complexes is mainly used for the detection of transition and heavy metal ions. After leaving the
separating column the eluate is mixed with a metallochromic reagent in a post-reactor; this reacts
with the analytes to form colored metal complexes which can be detected photometrically. A wide
range of azo-dyes can be used as the coloring agents [2,4]; these react with a large number of
metal cations. Transition metals and lanthanoides react with 4-(2-pyridylazo)-resorcinol (PAR)
(Figure 22) to form colored complexes. Lanthanoides and actinoides can be detected by using 2,7-
bis(2-arsenophenylazo)-1,8-dihydroxynaphthalene-3,6-disulfonic acid (Arsenazo III). Pyrocatechol-
3,5-disulfonic acid (Tiron) is suitable for the post-column derivatization of aluminum.
46 Metrohm
Monograph
PAR is mainly used for the detection of transition and heavy metal ions. With Fe, Co, Ni, Cu, Zn, Mn, Pb
and Cd PAR forms colored complexes; these absorb at wavelengths from 490 to 520 nm with absorption
coefficients up to 104 L mol –1 cm–1 and can be detected photometrically. The sensitivity of the method is
based on the fact that the extinction coefficients of the metal-PAR complexes are large compared with the
absorption coefficient of the reagent at the wavelengths used.
48 Metrohm
Monograph
possible as these could be destroyed by organic solvents. However, the Metrohm
Suppressor Module «MSM» is 100% solvent-resistant.
Cation analysis
In each analysis carried out the sample must be acidified (pH 2.5 … 3.5) with nitric acid
(approx. 100 µL 2 mol/L HNO3 to 100 mL sample), as otherwise no reproducible results
will be obtained for divalent cations.
Chemical grade
All the chemicals used should have at least the grades p.a. (analytical grade) or puriss.
(extra pure). Any standards used must be specially suitable for ion chromatography; for
example, sodium salts should be dissolved in water, not in acid.
Contamination sources
All solutions, samples, regenerating solutions, the water and the eluents used should be free of particles
as these will block the separating columns in the course of time (rise of column pressure). This is espe-
cially important when preparing the eluents as these flow through the column continuously (500 to 1000
mL per working day, whereas the sample throughput is approximately 0.5 mL).
Eluent degassing
In order to avoid bubble formation it is recommended that the water used for preparing the
eluent is degassed before the reagents are added to it. This can be done by applying a
vacuum from a water-jet or vacuum pump for about 10 minutes or by using an ultrasonic bath.
Environmental protection
One of the great advantages of ion chromatography is that usually aqueous media are used.
This is why the chemicals used in ion chromatography are to a large extent non-toxic and do
not pollute the environ-ment. However, if acids, bases, organic solvents or heavy metal
standards are used then care should be taken that these are disposed of properly after use.
Instrument shut-down
If the ion chromatograph is not to be used for a longer period of time (>1 week) then the
separating column should be removed and the ion chromatograph rinsed with methanol/water
(1:4). Care should be taken that all three chambers of the suppressor are rinsed.
Safety information
Protective goggles, protective clothing and, if necessary, protective gloves should be worn when carrying
out all the experiments. Be sure to observe the safety information (R/S-phrases) given for the chemicals.
Selecting a column
Most of the experiments described are carried out on very favorably priced columns – Metrosep
Anion Dual 1 for anions and Metrosep C2 for cations. These columns provide adequate separation
for all the experiments described. Of course, the Metrosep product range includes separating
columns which have a considerably better performance, but these are also more expensive.
50 Metrohm
Monograph
Water quality
In ion chromatography mainly aqueous media are used. This is why the water quality is crucial for good
chromatography results. If the water quality is unsatisfactory then it is certain that the results will also be
unsatisfactory. In addition there is the danger of damaging the instrument and separating column. The
deionized water used «aq. demin.» should therefore have a resistance greater than 18 MΩ and be
particle-free. It is recommended that the water is filtered through a 0.45 µm filter.
Instead of the Na+ and Cl– which were originally present in the sample the considerably
higher equivalent conductivity of H+ and Cl– is now measured and this also against a
lower background conductivity. Theo-retically a signal can be expected which is larger by
a factor of ten than for a measurement without chemical suppression. However, in
practice only an increase in sensitivity in the range of a factor of 2 to 4 is obtained.
In contrast to a linear calibration function obtained when working without chemical suppression,
when chemical suppression is used a quadratic calibration function is obtained; this means that a
multi-point calibration is necessary for its calculation. The linear concentration range is considerably
smaller (approx. 1/20 to 1/50) than when working without chemical suppression.
Learning content
• Principles of an IC system setup
• Differences between working with and without chemical suppression
• Determination of the different sensitivities of the two methods
Eluent preparation
Dissolve 1.33 g phthalic acid in 20 mL acetonitrile and a little water and make up to 1 L. Adjust
the pH to 4.1 by adding approx. 1 g TRIS (solid). Before addition of the reagents degas the
water used by applying a water-jet pump vacuum for 10 min or by using an ultrasonic bath.
52 Metrohm
Monograph
Chromatogram 1 Standard solution
Table 4 Components – Experiment 1a
Eluent preparation
Dissolve 265 mg sodium carbonate (anhydrous) and 201.5 mg sodium hydrogen
carbonate in 980 mL ultrapure water. Then add 20 mL acetone.
The system peak is significantly smaller here than in experiment 1a, chromatogram 1,
and can therefore only be discerned at highly sensitive settings of the conductivity scale.
54 Metrohm
Monograph
4.2.2 Experiment 2 – Capacity of separating columns
Cut-off peaks, shark-fin-shaped peaks, altered retention times for the same ions, peaks
with tailing or fronting – all these can have a common cause: overloading the column.
Each column has a finite number of ion exchange sites. Before the sample is injected these are
completely occupied by eluent ions. When the sample is injected, ion exchange starts to take place:
Eluent ions are exchanged for sample ions and sample ions for eluent ions. As the ion species differ
in their bonding constants they elute from the column at different speeds. The required result is the
separation of the mixture of substances and therefore the chromatogram.
This process only functions perfectly if the number of exchange sites is substantially higher
than the number of bonding places required by the sample. For example, a column with a
capacity of 1 meq (milli-equivalent) can bind a maximum of 1 mmol monovalent ions.
A second effect which has a negative influence on the separation is the fact that in principle each ion can
act as an eluent ion. If the column is overloaded then many eluent ions are replaced by sample ions. This
results in incalculable alterations to the column equilibria and the separation becomes worse.
The capacity of a column can be determined by completely occupying it with chloride ions.
After a rinsing step with deionized water the chloride ions are eluted with a carbonate eluent
and then quantified by either ion chromatography or argentometric titration.
In the following experiment the chloride concentration is increased until the column is
overloaded. Some effects associated with overloading the column are described below:
→ the retention times of the following ions become shorter,
→ the peak of the dominant ion is cut off,
→ the peak of the dominant ion tails,
→ the number of theoretical plates (TP) becomes smaller,
→ the area/height ratio becomes worse (at constant area, the peak height decreases) and
Learning content
• Explanation of the following chromatographic parameters: retention time, resolution,
area, number of plates, symmetry and area/height ratio
• Influence of a dominant main component on components with a much smaller
concentration: alter-ations to the above-mentioned parameters
Practical Ion Chromatography
55
Table 7 Parameters – Experiment 2
Eluent preparation
Dissolve 265 mg sodium carbonate (anhydrous) and 201.5 mg sodium hydrogen
carbonate in 980 mL ultrapure water and then add 20 mL acetone.
56 Metrohm
Monograph
Chromatogram 3b Standard solution with a very high chloride concentration; enlarged part of chromatogram 3a
Remarks
Sodium chloride is weighed in. The retention times of the peaks as well as the values for
the number of plates and area may vary depending on the quantity of NaCl weighed in.
The huge chloride peak interferes with the evaluation of the following peaks. Correct identification of
the individual peaks is only possible by spiking the sample solution with the respective anions.
Learning content
• Comparison of eluents with monovalent and divalent anions
• Comparison of sodium hydroxide and carbonate/hydrogen carbonate eluents
• Influence of the eluent pH on the retention of phosphate
Table 8 Parameters – Experiments 3a to 3d
58 Metrohm
Monograph
Experiment 3a – 2.5 mmol/L Na2CO3 / 2.4 mmol/L
60 Metrohm
Monograph
Experiment 3c – 4 mmol/L Na2CO3 / 1 mmol/L NaOH
Eluent preparation
Dissolve 424 mg sodium carbonate (anhydrous) and 40 mg NaOH in 1 L ultrapure water.
Eluent preparation
Dissolve 106 mg sodium carbonate (anhydrous) and 160 mg NaOH in 1 L ultrapure water.
62 Metrohm
Monograph
4.2.4 Experiment 4 – Calibration, detection and determination limits in ion chromatography
Important parameters for analytical determination methods are the linear range, detection limit and deter-
mination limit. The mathematical methods for these are laid down in standards, e.g. in DIN 32645.
If chromatograms are recorded with a conductivity detector then the peak area is usually used for
evalua-tion. The peak area is proportional to the amount of substance. If the peak area is plotted
against the concentration then the calibration function is obtained. It is linear for measurements
without chemical suppression. As a first approximation it is a quadratic function for measurements
with chemical suppres-sion. Evaluation programs calculate the calibration functions automatically.
Evaluation using peak height is used preferentially for strongly tailing peaks or for
insufficiently separated peaks with strongly differing area/height ratios as in those cases
the area-based evaluation leads to large errors.
The detection limit is the minimum concentration of an analyte which can still be
detected with known statistical certainty. The theoretically smallest concentration which
can just be distinguished from a blank value is calculated.
There are two methods for calculating the detection limit:
– Blank value method:
The blank sample should be a sample which does not contain the ion to be determined but
produces a signal at the same location as the sample ion. The repeated measurement of a blank
sample gives for a concentration «0» (X value) measured values (Y values) whose mode (average
maximal frequency) is known as the blank value. By using the calibration curve the maximum Y
value is assigned to a concentra-tion value on the X axis; this is the detection limit.
– Calibration curve method:
This method is used when it is not possible to determine a blank value because the ion under investigation
cannot be detected in the sample. In the calibration curve method multiple measurements are carried out
on different concentrations of the ion. A range of confidence is then obtained from the standard deviation.
In this way the concentration «0» corresponds to a particular Y interval. The calibration function is used to
assign the Y interval to a concentration interval whose maximum value is the detection limit.
The signal/noise ratio is often used to determine the detection limit. For example, the
detection limit is defined as being the analyte concentration at which the measuring
signal is 3 times, 5 times or 7 times the baseline noise.
The determination limit is reached when the measuring error undercuts a particular value when compared
with the analytical value, e.g. 1/3. Only then will a numerical value be given in the analytical report, as
otherwise the measuring error is regarded as being too large in comparison to the analytical value. As an
estimation it can be said that the determination limit is higher than the detection limit by a factor of three.
Learning content
• What does calibration mean?
• Comparison of a single-point and a multi-point calibration – estimation of error
• Determining the system noise
• Estimating the detection limit
Eluent preparation
Dissolve 265 mg sodium carbonate (anhydrous) and 201.5 mg sodium hydrogen
carbonate in 980 mL ultrapure water and then add 20 mL acetone.
64 Metrohm
Monograph
Table 14 Components – Experiment 4a
Eluent preparation
Dissolve 1.33 g phthalic acid in 20 mL acetonitrile and a little water and make up to 1 L.
Adjust the pH to 4.1 by adding TRIS (solid).
66 Metrohm
Monograph
4.2.5 Experiment 5 – Altering the selectivity with the aid of crown ethers (18 Crown-6)
The retention times of cations can be altered by adding complexing agents to the eluents. The
complexing agent acts as a ligand, the analyte cation is included as the central metal ion. The
more selective a ligand with respect to a central metal ion, the stronger the influence on the
retention time. In an ideal case the retention times of the other cations will only alter slightly.
Complexing agents are used to obtain better separation of alkali metal ions. The addition of the crown ether
18 Crown-6 to the eluent leads to a better separation of Na +, NH4+ and K+. For example, the crown ether
can be added to the eluent to improve the separation between Na + and NH4+ when traces of NH4+ have to
be determined in natural waters. The increase in the retention time of K + is particularly dramatic. This can
be explained by the formation of the complex from K + and dibenzo-18-crown-6-ether (18 Crown-6). K+ fits
exactly in the «cage» of the ether. It is complexed via the electron pairs of the oxygen atoms. After
complexation, a considerably larger molecule with the same charge is separated. This means that the
retention time of the potassium is increased as a result of steric hindrance.
The name 18 Crown-6 indicates that the ring
system consists of 18 atoms of which 6 are oxygen
atoms. Crown ethers not only play an important
role in ion chro-matography, but are also used as
the ion-selective phase in potassium electrodes.
Learning content
• Effect of a very selective complexing agent on the
retention times
Figure 26 Potassium – 18 Crown-6 • Explanation of the effect in comparison with Experi-
ment 6
Table 17 Parameters – Experiments 5a and 5b
Eluent preparation
Dissolve 600 mg tartaric acid and 125 mg dipicolinic acid in 100 mL ultrapure water
under heating and then make up to 1 L with ultrapure water.
68 Metrohm
Monograph
Experiment 5b – Eluent with crown ether
Eluent preparation
Dissolve 600 mg tartaric acid and 125 mg dipicolinic acid in 100 mL ultrapure water
under heating, add 132 mg crown ether and make up to 1 L with ultrapure water.
Me 2+ dipicolin complex
This means that, depending on the pH, the complex formed has either a double positive
charge, a single positive charge, or is uncharged.
The primary separation criterion on a cation exchanger column is the charge of the ions to be separated.
Uncharged complexes are not retarded whereas complexes with a 3-fold positive charge are bound very
strongly. As a result of the complex formation constant and the set pH the complex has a certain average
equilibrium charge. This equilibrium charge determines the retention time. This is the reason why divalent
metal ions can be accelerated by the addition of dipicolinic acid within a particular pH range.
There is no influence on the retention times of monovalent metal ions, which do not form
a complex with dipicolinic acid.
Learning content
• Influence of the complex formation constant on the retention time – comparison of zinc and calcium
70 Metrohm
Monograph
Table 20 Parameters – Experiments 6a to 6d
Eluent preparation
Dissolve 600 mg tartaric acid in 1L ultrapure water.
72 Metrohm
Monograph
Experiment 6b – 4 mmol/L tartaric acid + 0.1 mmol/L
dipicolinic acid Eluent preparation
Dissolve 600 mg tartaric acid and 17 mg dipicolinic acid under heating in 100 mL
ultrapure water and make up to 1L.
Chromatogram 13 Standard solution – Eluent 4 mmol/L tartaric acid + 0.1 mmol/L dipicolinic acid
Chromatogram 14 Standard solution – Eluent 4 mmol/L tartaric acid + 0.25 mmol/L dipicolinic acid
74 Metrohm
Monograph
Experiment 6d – 4 mmol/L tartaric acid + 0.75 mmol/L
dipicolinic acid Eluent preparation
Dissolve 600 mg tartaric acid and 125 mg dipicolinic acid under heating in 100 mL
ultrapure water and make up to 1L.
Chromatogram 15 Standard solution – Eluent 4 mmol/L tartaric acid + 0.75 mmol/L dipicolinic acid
Remarks
All solutions must be stored in plastic vessels. For a correct determination of sodium all contact with
glass must be avoided. The pH of the standard and sample solutions must be between 2.5 and 3.5.
After changing the eluent allow the system to run until the baseline is constant.
Chromatograms 14 and 15: Zinc is complexed by the dipicolinic acid and elutes in the front peak.
Figure 27 Preconcentration technique: on the left «fill» position, on the right «inject» position
76 Metrohm
Monograph
Learning content
• What is sample preparation?
• Where is the interface between ion chromatography and sample preparation?
• What is the advantage of sample preconcentration?
• What are the limits of the method?
• What is the reason for using alkaline eluents with suppression?
• What is the effect of CO2/carbonate in the sample?
Eluent preparation
Dissolve 265 mg sodium carbonate (anhydrous) and 201.5 mg sodium hydrogen
carbonate in 980 mL ultrapure water and add 20 mL acetone.
78 Metrohm
Monograph
Chromatogram 17 Ultrapure water
Remarks
Thoroughly rinse all vessels, syringes and the system several times; use plastic vessels.
Depending on the degree of pollution various methods are used for the treatment of drinking water.
80 Metrohm
Monograph
Disinfection is always necessary when the water is not pathogen-free. Chlorine, ozone,
chlorine dioxide and UV irradiation are all used.
Preventative chlorination before release into the drinking water pipeline in order to
prevent growth of microorganisms on the way to the consumer.
Learning content
• Investigation of the «Number 1» foodstuff
• Checking the information given on bottles of mineral water
Eluent preparation
Dissolve 265 mg sodium carbonate (anhydrous) and 201.5 mg sodium hydrogen
carbonate in 980 mL ultrapure water and add 20 mL acetone.
82 Metrohm
Monograph
Chromatogram 19 Mineral water without carbon dioxide
Remarks
Mineral water containing carbon dioxide must be degassed before the measurement – why?
Learning content
• Analysis of foodstuffs
• Influence of the matrix on chromatography
• Sample preparation
• Sample preconcentration and matrix elimination
84 Metrohm
Monograph
Eluent preparation
Dissolve 265 mg sodium carbonate (anhydrous) and 201.5 mg sodium hydrogen
carbonate in 980 mL ultrapure water. Then add 20 mL acetone.
86 Metrohm
Monograph
Experiment 9b – Determination of anions in whiskey
Remarks
Peaks 3, 4 and 10 not evaluated.
88 Metrohm
Monograph
Experiment 9c – Determination of anions in ethanol with preconcentration
and matrix elimination
Remarks
Thoroughly rinse all vessels, syringes and the system several times; use plastic vessels.
90 Metrohm
Monograph
4.3.3 Experiment 10 – Anions in lettuce
Nitrate enters the human body via drinking water, but also by eating vegetables, salad, etc.
The WHO (World Health Organization) recommends that the daily nitrate consumption should
not exceed 220 mg per person. In Germany the average daily consumption is about 130 mg
per person per day. Approx. 5% of this originates from meat and sausages; the remainder is
split about 50:50 between drinking water and veg-etables, salad, etc.
Nitrate can have an adverse effect on human health in several different ways. Nitrate itself is relatively
non-toxic. Only when larger amounts are consumed can this lead to inflammation of the gastro-intestinal
tract. However, under certain conditions, particularly in the presence of bacteria such as are present in the
human mouth, nitrate is reduced to nitrite by the enzyme nitrate reductase:
Nitrite is able to convert hemoglobin, the red blood colorant, to methemoglobin (Fe 2+ is oxidized to Fe3+). In
contrast to hemoglobin, methemoglobin cannot transport oxygen in the blood. In adult human beings this
damage can be remedied by metabolic reactions. However, the metabolism of babies up to 5 months old
is not yet able to do this. This results in a lack of oxygen in the baby's blood and the skin turns bluish. This
cyanosis or methemoglobinemia can sometimes result in death.
In the acidic surroundings of the stomach nitrite can react with various secondary
amines (two H atoms of the ammonium molecule are replaced by alkyl residues), which
enter the human body via medicines and foodstuffs, to form nitrosamines.
Nitrosamines are among the most carcinogenic substances known. They are formed
within the organism from non-carcinogenic compounds.
Learning content
• Analysis of foodstuffs
• Checking foodstuffs for nitrite and nitrate in particular
Eluent preparation
Dissolve 265 mg sodium carbonate (anhydrous) and 201.5 mg sodium hydrogen
carbonate in 980 mL ultrapure water. Then add 20 mL acetone.
Sample preparation
Chop up lettuce, disperse, add ultrapure water (ratio 1:100) and filter.
92 Metrohm
Monograph
Chromatogram 26 Standard solution for the analysis of lettuce
Remarks
Peaks 3, 9 and 10 not evaluated.
94 Metrohm
Monograph
4.3.4 Experiment 11 – Phosphoric acid in cola drinks
Coca-Cola is a non-alcoholic beverage which contains carbon dioxide and caffeine. It was first
made in 1885 by the American pharmacist Pemberton. Among its ingredients are extracts of
cola nuts, bitter oranges, locust beans, ginger essence, 12% sugar, 0.28% phosphoric acid
(pH 2.7), caramel color and carbon dioxide. The caffeine content is 16 mg/100 mL.
The determination of the phosphoric acid content in the cola drinks is of special
significance as the local cola bottlers get the concentrate and the bottling is controlled
via the phosphate content, requiring a highly accurate determination.
Many cola drinks are available on the market (Pepsi Cola, Club Cola, River Cola, Afri
Cola, etc.); the composition of these may differ from Coca-Cola.
Phosphoric acid is a tribasic acid (pK A1= 2.161; pKA2= 7.207; pKA3= 12.325). The amounts of
the different species vary with the pH as can be seen in the following diagram:
A mixture of primary and secondary phosphate is frequently used as a buffer solution which
buffers in the pH range 6 ... 8 (90% H2PO4– + 10% HPO42– to 10% H2PO4– + 90% HPO42–).
See also the Henderson-Hasselbalch equation (buffer equation)
Eluent preparation
Dissolve 265 mg sodium carbonate (anhydrous) and 201.5 mg sodium hydrogen
carbonate in 980 mL ultrapure water. Then add 20 mL acetone.
96 Metrohm
Monograph
Chromatogram 28 Standard solution for the analysis of cola
98 Metrohm
Monograph
Experiment 11b – Analysis of low-calorie cola (1:10 dilution)
Remarks
Peaks 6, 7 and 8 not evaluated.
Remarks
The cola must be degassed. In low-calorie cola the sweetener cyclamate lies under the phosphate peak.
100 Metrohm
Monograph
4.3.5 Experiment 12 – Organic acids in wine
According to the German wine law, wine is a product which is only obtained by the
complete or partial alcoholic fermentation of fresh or mashed grapes or from grape juice.
Grape juice contains about 12 … 25% carbohydrates (glucose, fructose) and 0.9 …
1.5% acids; the most important are L-(+) tartaric acid (2R, 3R) and malic acid, but citric
acid, ketoglutaric acid, succinic acid and lactic acid may also be present.
An important criterion for assessing the grape juice is the Oechsle degree (Oe°); the higher it is the more
sugar the juice contains. It indicates the number of grams by which 1 L juice at 20 °C is heavier than 1 L
distilled water. For example, juice with a density of 1.115 kg/L (115 g more than 1 L water) has 115
°Oechsle. From the degrees Oechsle a simple calculation permits determination of the sugar and alcohol
content. 1.7 g sugar produce 1 mL (0.794 g) ethanol. At an alcohol volume fraction of 12 … 15% fermen-
tation comes to a standstill as the yeasts are killed off by the alcohol they have produced.
The aroma of the wine is made up of 600 to 800 components: hydrocarbons, alcohols,
aldehydes, ke-tones, acids, esters, lactones, ethers, phenols and many more.
Of analytical interest are the 2R, 3R-tartaric acid, malic acid, citric acid, lactic acid and
succinic acid. The total acid content (calculated as tartaric acid) is usually between 5.5
and 8.5 g/L. Acetic acid, propionic acid, higher fatty acids and abnormal amounts of
lactic acid occur in "spoiled" wine and are chiefly pro-duced by microorganisms.
L-(+)-tartaric acid
(2R, 3R)-form
The weak organic acids are determined by ion exclusion chromatography (see chapters
3.3.4, 3.6.7 and 3.7.2.3).
Remarks
– = not detectable ± = present in traces (approx. 5 mg/L), not evaluated
+ = traces 3-M-2,3-DHBS = 3-methyl-2,3-dihydroxybutyric acid
Learning content
• Analysis of foodstuffs
• Ion exclusion chromatography
• Which ions can be determined by ion exclusion chromatography?
Eluent preparation
Mix together 0.5 mmol H2SO4 (850 mL) and acetone (150 mL).
Sample preparation
Dilute sample 1:100, filter wine through 0.45 µm filter.
104 Metrohm
Monograph
Chromatogram 33 Determination of organic acids in white wine (1:100 dilution)
Table 50 Components – Experiment 12a – White wine
Remarks
Peaks 3, 5, 6 not evaluated.
Remarks
Peaks 3, 5, 6 were not identified; citrate and malate could not be quantified correctly.
106 Metrohm
Monograph
4.3.6 Experiment 13 – Determination of contaminants in borate – chloride and
sulfate in borax solutions
Borax (disodium tetraborate, Na2B4O7·10 H2O, Na2[B4O5(OH)4]·8H2O) has the following structure:
When it melts borax can dissolve many metal oxides with the formation of characteristic colors. These
«borax pearls» are well-known in practical inorganic chemistry. Borax is also used in the manufacture of
glass, pottery glazes, porcelain and as a flux in brazing. At 100 °C borax loses 5 water molecules and
becomes the pentahydrate «jeweler's borax». If a soldering flux is added to it then surface contaminants –
mainly oxide layers – are destroyed. These would otherwise affect the formation of an alloy between the
solder (itself an alloy of silver, copper and tin) and the basic material.
Sodium tetraborate (borax) solutions are used in the internal cooling circuit of nuclear power
plants as neutron absorbers. The purity of the material is extremely important as traces of
chloride and sulfate cause corrosion in the piping, which must be avoided by all means.
Boric acid H3BO3 or B(OH)3 is a very weak monobasic acid whose pK A value of 9.25
corresponds approxi-mately to that of HCN. Boric acid is not an H+ donor (Brönsted acid
definition), but rather an OH– acceptor (Lewis acid definition).
Ion chromatography without chemical suppression uses an acidic eluent with pH = 4.1 for the
determina-tion of the contaminants in borate. At this pH borate is present almost entirely as boric
acid B(OH)3. In contrast to the negatively charged anions, boric acid does not interact with the
stationary phase of the separating column. This means that it elutes in the dead volume.
However, with chemical suppression the weakly alkaline carbonate/hydrogen carbonate
eluent is used instead. Although it is now possible to separate borate as an anion, the
detection does not function. The cause of this can be found in the suppressor, which
exchanges all the Na+ ions by H+ ions. As a result, the weak boric acid is formed.
The large amount of boric acid disturbs the chromatogram, making the determination of the anions
almost impossible. Matrix elimination (see chapter 4.3.2) is the best possibility to solve the problem.
In power plants, traces of anions must often be determined in water samples containing
2 to 4% boric acid. In these cases, the matrix elimination technique with sample
preconcentration is applied (see chromatogram 37).
Learning content
• Strong and weak acids
• Comparison between measurements with electronic and chemical suppression
• Sample preparation
• Sample preconcentration and matrix elimination
Eluent preparation
Dissolve 1.33 g phthalic acid in 20 mL acetonitrile and a little water, add the solution to degassed
water and make up to 1 L. Adjust the pH to 4.1 by the addition of solid TRIS (approx. 1 g).
Chromatogram 35 Borate spiked with 1 ppm chloride and 1 ppm sulfate – without chemical suppression
108 Metrohm
Monograph
Table 53 Components – Experiment 13a
Eluent preparation
Dissolve 265 mg sodium carbonate (anhydrous) and 201.5 mg sodium hydrogen
carbonate in 980 mL ultrapure water. Then add 20 mL acetone.
110 Metrohm
Monograph
Eluent preparation
Dissolve 583 mg sodium carbonate (anhydrous) and 67.2 mg sodium hydrogen
carbonate in 1 L ultrapure water.
Chromatogram 37 Determination of trace anions in 4% boric acid with preconcentration and matrix elimination
Apart from the oxidation of organic substances, ammonium can also be oxidized to
nitrate (nitrification) in suitably designed plants.
In oxygen-free basins bacteria utilize the nitrate oxygen for the oxidation of organic
substances (denitrifica-tion).
Phosphate – a plant nutrient like NH4+ and NO3– – can be precipitated, e.g. by the addition of
Fe3+ salt solutions. Apart from the so-called sum parameters – biochemical oxygen demand
(BOD), chemical oxygen demand (COD), total organic carbon (TOC) – which are a measure
of the organic load of the wastewater or water in general, the analysis of NH 4+, NO3–, PO43– is
also important. Cl–, SO42– are normally only analyzed under special circumstances.
In communal treatment plants with more than 100 000 so-called population equivalents (1 population
equivalent is the sewage flow per head) the following limits have been laid down in Germany:
BOD 15 mg/L
COD 75 mg/L
NH4-N 10 mg/L*
NTot 18 mg/L* (sum of NH4-N, NO2-N, NO3-N)
PO4-PTot 1 mg/L
* Only applies at a wastewater temperature ∞ 12 °C, as nitrification is greatly influenced by the temperature.
NH4-N, NO2-N, NO3-N means that all the values of the nitrogen contained in the ions is
taken into consid-eration; the same applies for PO4-P (P content).
Learning content
• Environmental analysis
• Sample preparation techniques
• Analysis of mixtures of substances with large differences in concentration – dynamic
concentration range
112 Metrohm
Monograph
Table 58 Parameters – Experiment 14a and 14b
Eluent preparation
Dissolve 265 mg sodium carbonate (anhydrous) and 201.5 mg sodium hydrogen
carbonate in 980 mL ultrapure water and add 20 mL acetone.
Sample preparation
Filtering of the samples is absolutely essential. Use, for example, Minisart (Sartorius) or
one-way filter holders (Schleicher & Schüll) with a pore size of 0.45 µm or less. Even
seemingly clear solutions can contain very fine particles that damage the column.
114 Metrohm
Monograph
Chromatogram 39 Analysis of a treatment plant inflow
Remarks
Peaks 2, 3, 4 and 8 not evaluated.
116 Metrohm
Monograph
Chromatogram 41 Analysis of a treatment plant outflow
Remarks
Peak 3 not evaluated.
Sodium monofluorophosphate
• Surfactants (e.g. sodium lauryl sulfate) – lower the surface tension and help to distribute
the tooth-paste evenly. They increase the cleaning effects, particularly in locations which
are difficult to reach with the toothbrush. Other ingredients of toothpastes are e.g.:
• Moisteners (e.g. glycerol, sorbite, polyethylene glycol) – improve the low-temperature
stability and prevent drying out.
• Binders and thickeners (e.g. sodium carboxymethylcellulose) – prevent separation
into liquid and solid phases.
• Sweeteners (e.g. sodium saccharin) – improve the taste of the toothpaste.
• Preservatives (e.g. 4-hydroxybenzoic acid) – are used to protect the toothpaste
against bacterial decomposition.
• Aromatic substances – are added to increase both the acceptance of the toothpaste
and also hygiene consciousness. These include peppermint oil, menthol, aniseed oil,
eucalyptus oil, aromatic oils, citrus oil.
Analysis of toothpaste by ion chromatography determines the fluoride originating from
sodium fluoride together with the sodium monofluorophosphate.
Remarks
Citrate elutes from the Metrosep Anion Dual 1 (3 x 150 mm) only after approx. 90 min.
In order to check whether citrate is present in the toothpaste, Experiment 15b uses the Metrosep Anion
Dual 1 precolumn cartridge (6.1006.030) instead of the Metrosep Anion Dual 1 separating column.
118 Metrohm
Monograph
Learning content
• Quality control
• Sample preparation
Eluent preparation
Dissolve 265 mg sodium carbonate (anhydrous) and 201.5 mg sodium hydrogen
carbonate in 980 mL ultrapure water. Then add 20 mL acetone.
Sample preparation
Dissolve 1 g toothpaste in 19 mL ultrapure water and then filter it through a 0.45 µm filter.
120 Metrohm
Monograph
Pressure 3 MPa
Analysis time 25 min
Loop 20 µL
Suppressor Regenerating agent: 50 mmol/L H2SO4, ultrapure water
Autostep with Fill
Polarity +
Eluent preparation
Dissolve 265 mg sodium carbonate (anhydrous) and 201.5 mg sodium hydrogen
carbonate in 980 mL ultrapure water. Then add 20 mL acetone.
Sample preparation
Dissolve1 g toothpaste in ultrapure water (1:20 dilution) and then filter it through a 0.45 µm filter.
In the manufacture of sugar from sugar beet the beet are first washed and chopped up; they are
then extracted with water in so-called diffusion units in counterflow. The soluble constituents –
sugar, salts, acids, proteins, pectins – are dissolved. The major share of the non-sugar constituents
is precipitated out by the addition of burnt lime (CaO). Carbon dioxide (CO 2) is then used to
precipitate the excess calcium hydroxide as CaCO 3. After filtration the sugar solution is
concentrated in multi-stage evaporators to form «thick juice», filtered again and concentrated further
until the sugar separates out as «white massecuite». The sugar is separated off in centrifuges and,
after purification (recrystallization), snow-white crystal sugar with a purity of 99.95% is obtained. The
centrifuge run-off of the last stage is a brown-colored syrup – molasses.
Brown sugar is less intensively purified and may be colored by the addition of molasses.
Apart from organic contaminants, it contains alkali metal and alkaline earth ions as well
as anions; it also has a different taste to white crystal sugar.
The organic contaminants can be separated off with an RP column before the anions
are determined. Learning content
• Analysis of foodstuffs
• Process control
• Checking foodstuffs containing large amounts of sugar, e.g. honey
122 Metrohm
Monograph
Table 67 Parameters – Experiments 16a and 16b
Eluent preparation
Dissolve 265 mg sodium carbonate (anhydrous) and 201.5 mg sodium hydrogen
carbonate in 980 mL ultrapure water. Then add 20 mL acetone.
Sample preparation
Dissolve the sugar in ultrapure water (1:10 dilution) and filter through a 0.45 µm filter.
124 Metrohm
Monograph
Chromatogram 45 Analysis of white sugar (1:10 dilution)
126 Metrohm
Monograph
Chromatogram 47 Analysis of brown sugar (1:10 dilution)
Remarks
With brown sugar, contaminants may elute at later retention times; this could interfere
with the following chromatograms if the run time is too short. This is why the white sugar
should be measured first, followed by the brown sugar.
128 Metrohm
Monograph
Learning content
• Process control
• Sample preparation
• Sample preconcentration and matrix elimination
• Selectivity of different columns
Eluent preparation
Dissolve 265 mg sodium carbonate (anhydrous) and 201.5 mg sodium hydrogen
carbonate in 980 mL ultrapure water. Then add 20 mL acetone.
Sample introduction
Rinse the system thoroughly with ultrapure water, then pass 1 mL sample over the
preconcentration column and rinse with 1 mL ultrapure water.
130 Metrohm
Monograph
Experiment 17a – Analysis of hydrogen peroxide without matrix elimination
Remarks
Rinse vessels, syringes and system thoroughly several times. Use plastic vessels.
Protective goggles must be worn! Peaks 1, 4 and 9 not evaluated.
132 Metrohm
Monograph
Experiment 17c – Analysis of hydrogen peroxide with matrix elimination on a
high-perfor-mance column
Eluent preparation
Dissolve 339 mg sodium carbonate (anhydrous) and 84 mg sodium hydrogen carbonate
in 1 L ultrapure water.
Sample introduction
Rinse the system thoroughly with ultrapure water, then pass 1 mL sample over the
preconcentration column and rinse with 1 mL ultrapure water.
Remarks
Peaks 1, 2, 5. 9, 10, 11 and 12 not evaluated.
134 Metrohm
Monograph
4.4 Experiments for determining cations
4.4.1 Experiment 18 – Alkali metals and alkaline earth metals in drinking
water and mineral water
Apart from the dissolved gases O 2, N2 and CO 2, natural water also contains salts which are mainly
leached out of the soil and rocks to enter the ground water. Another ingredient of natural water are polar
organic compounds originating, for example, from the humus layers of the soil. Contamination by
wastewater is also a possible source of various salts and organic compounds. The most important salts
are the chlo-rides, sulfates and hydrogen carbonates of sodium, calcium and magnesium.
An important parameter of drinking water, both for its use as a foodstuff and its use in
washing processes or industrial processes, is the so-called water hardness.
The total hardness of the water is understood as being the sum of the molar concentrations
(mmol/L) of the calcium and magnesium ions. That part of the hardness which can be removed by
boiling is known as the carbonate hardness (previously temporary hardness). The residual
hardness is caused by sulfate and chloride ions, whose Ca and Mg salts cannot be precipitated out
by boiling.
With the classical soaps – sodium salts of fatty acids – calcium and magnesium ions form insoluble
calcium and magnesium compounds. These are both ineffective in washing and are also deposited
on textiles as a «gray frosting». Modern washing agents contain alkyl sulfates which do not form
insoluble calcium and magnesium compounds. The zeolites which are also added to these washing
agents function as ion exchangers and bind Ca2+ and Mg2+ ions. This effectively prevents the
precipitation of insoluble CaCO3 and basic magnesium carbonate on heating.
The drinking water regulations, which are the implementation of an EC drinking water directive (80/778/
EEC), contain the following values for cations:
An increased intake of sodium ions by human beings causes high blood pressure, among other things.
The daily Na+ requirement of 1 g is greatly exceeded today (3 to 7 g).
In all analyses carried out the samples must be acidified by the addition of nitric acid (pH
2.5 to 3.5), as otherwise no reproducible results will be obtained for divalent cations.
If the sodium content of the sample ist substantially higher than that of the standard
solution, the latter’s sodium content should be adapted to that of the sample.
Learning content
• Checking the «Number 1» foodstuff
• Checking the information given on bottles of mineral water
• Why is it necessary to acidify the samples?
Eluent preparation
Dissolve 600 mg tartaric acid and 125 mg dipicolinic acid in 100 mL ultrapure water
under heating and make up to 1 L with ultrapure water.
136 Metrohm
Monograph
Table 78 Components – Experiment 18 – Standard
Remarks
All solutions must be stored in plastic vessels. All contact with glass must be avoided for a correct sodium
determination. The pH value of the standard and sample solutions should be between 2.5 and 3.5.
138 Metrohm
Monograph
4.4.2 Experiment 19 – Determination of transition metals
Complexing agents are added to the mobile phase in order to separate the transition metal ions. They
reduce the charge density and improve the selectivity of the separation, which is relatively low for transi-
tion metal ions having the same charge. Apart from the equilibrium between exchanger resin and analyte
ion, if complexing agents are added there is a further equilibrium, namely between metal ion and
complexing agent. This equilibrium is also influenced by the pH of the solution if various dissociation steps
of the complexing agent are possible or if the metal ions form hydroxy complexes.
Weak organic acids are often used as complexing agents, e.g. tartaric acid, citric acid, oxalic acid, pyri-
dine-2,6-dicarboxylic acid (dipicolinic acid – see chapter 4.2.7 – Experiment 6 – Altering the selectivity by
using complexing agents). The number of ligands L surrounding the central metal ion (Me) can vary, e.g.
MeL, MeL2, MeL 3. The resulting charge can also differ. Depending on the charge of the metal ion, the
ligands and their number, either cationic, neutral or anionic complexes can be present. This means that,
depending on their charge, the complexes can be separated on cation exchangers or anion exchangers.
Theoretically the use of exchangers containing both cation and anion exchanger groups is possible.
The separation can be influenced and optimized by varying the pH value and the complexing
agents – two or more different complexing agents can be added to the mobile phase. In part
the influences are contrary and difficult to predict. Some general information is given below:
• Different stabilities (complex formation constants) influence the separation.
• The addition of an organic solvent, e.g. acetone, to the eluent also influences the
separation. Lipophilic ions elute more quickly after the addition.
• In cation exchange chromatography, ethylenediamine has proven itself as an eluting
agent. Under normal conditions it is present as a cation carrying two protons.
• An increase of the pH results in a reduction of the retention time by reducing the total charge.
• An increase in the ligand concentration in the mobile phase, apart from displacing the
complexing equilibrium, also increases the concentration of the counter-ion H+ or Na+
and accelerates the dis-placement of the complex from the negatively charged
exchange sites, i.e. the separation performance is reduced.
• An increase in the ligand concentration causes an increase in the retention time as the coordination
number of the complex increases, but the free anionic ligands also promote the elution of the anionic
metal complex, so that two opposing effects exist with regard to the retention time.
Detection of the complexes can be carried out either by conductivity or photometrically
(post-column derivatization).
For conductivity detection only detection without chemical suppression can be used as
insoluble hydrox-ides would be formed by the suppressor reaction and the only eluent
anions available are carbonate or hydroxide.
Photometric detection requires post-column derivatization in which an additional, chromophoric
complexing agent displaces the one originally present (e.g. tartaric acid, oxalic acid...). The newly
formed and UV- or VIS-absorbing complex is detected, whereby the absorption maximum of the
free ligands must not be in the region of the maximum of the complex. As, for most complexing
dyestuffs, the maxima of the indi-vidual transition-metal complexes lie in different wave-length
regions, the detection limits of the individual metals differ greatly.
140 Metrohm
Monograph
Experiment 19a – Determining transition metals with an eluent containing tartaric
acid, citric acid, ethylenediamine and acetone
Eluent preparation
Dissolve 600 mg tartaric acid and 105 mg citric acid in ultrapure water. Add 200 µL
ethylenediamine and 50 mL acetone and make up to 1 L.
142 Metrohm
Monograph
Chromatogram 57 Analysis of tap water (Eluent a – 1:10 dilution)
Eluent preparation
Dissolve 315 mg oxalic acid in ultrapure water, add 50 mL acetone and make up to 1 L
with ultrapure water. Adjust to pH = 4 with ethylenediamine (approx. 120 µL).
144 Metrohm
Monograph
Chromatogram 59 Analysis of tap water (Eluent b, 1:10 dilution)
Remarks
Calcium and manganese coelute if Eluent a is used. With Eluent b calcium and manganese
are separated; nickel, zinc and cobalt are strongly complexed and elute in the water peak.
Apart from its use as a technical adsorption agent, silica gel is used in both gas
chromatography and high pressure liquid chromatography.
In gas chromatography originally only packed columns were used – they still are used
to a small extent today. A liquid phase, e.g. silicone oil, is applied to a particulate carrier
such as silica gel. The number of theoretical plates of such columns is low compared
with those of the capillary columns which are mainly used today.
In high pressure liquid chromatography (HPLC) silica gel is used as a stationary phase or carrier material for
adsorption, reversed phase (RP), normal phase and ion chromatography. For some of these applica-tions the
silica gel surfaces have to be modified with functional groups (e.g. C18 or ion exchanger groups).
In adsorption chromatography silica gel is also used as a stationary phase or in addition to
Al2O3. The chromatographic separation is based on dipole-induced dipole linkage, dipole-
dipole linkage, hydrogen bridge linkage and π-complex linkage. Adsorption chromatography is
used for the separation of non-polar substances which are difficult to dissolve in water.
In RP chromatography the free silanol groups of the silica gel are hydrophobicized by
the chemical bond-ing of longer alkyl chains (e.g. octyl or octadecyl groups):
In this conversion free residual silanol groups remain; these can adsorb polar compounds and cause peak
tailing. These residual silanol groups can be deactivated by trimethylchlorosilane – «end capping».
In normal phase chromatography the polar residues are chemically bonded to the silanol groups, e.g.:
A less polar or nonpolar mobile phase is used with this polar stationary phase for chromatographic sepa-
ration. In RP chromatography, which is used in approx. 75% of all HPLC applications, the polarities are
reversed and a nonpolar stationary phase is used with a polar mobile phase. This means that the name
reversed phase has a historical basis as this method was only developed later.
146 Metrohm
Monograph
In ion chromatography silica gel to which sulfonic acid groups have been bonded is
used as a cation exchanger for the analysis of alkali ions.
Contamination of silica gel by metal ions is important in chromatography as these can cause non-
specific interactions or blank values. A strongly acidic cation exchanger is used for the analysis of
these contami-nants; monovalent cations elute in the dead volume and only the divalent cations are
separated. Fe(III) must be reduced by ascorbic acid to Fe(II) before the analysis.
Learning content
• Quality control
• Additional experiment: spiking the extract with Fe(III); determination with and without
the addition of ascorbic acid (vitamin C)
• HPLC separating columns based on silica gel
Eluent preparation
Dissolve 600 mg tartaric acid and 105 mg citric acid in ultrapure water. Add 200 µL
ethylenediamine and 50 mL acetone and make up to 1 L with ultrapure water.
148 Metrohm
Monograph
Chromatogram 61 Determination of cations in silica gel
Aminoethanol Iminodiethanol
Nitrilotriethanol
All three substances are used for removing CO2 and H2S from mixtures of gases, e.g. for
removing carbon dioxide from synthesizer gas and in the synthesis of ammonia as well
as for the removal of hydrogen sulfide from refinery gases.
At a higher temperature CO2 is again released and the ethanolamine NR3 is returned to
the absorption process.
The H2S released again at the higher temperature is oxidized to liquid sulfur in «Claus
units»; this is converted to sulfuric acid via SO2 and SO3.
Monoethanolamine (MAK = maximum allowable workplace concentration = 3 ppm) is
used in surfac-tants as a fatty acid ester. Diethanolamine is used in furniture and floor
care products, shoe polish and lubricants.
Triethanolamine forms triethanolamine soaps with fatty acids, e.g. with stearic acid
C18H35COOH. These are easily soluble in water and mineral oils and are therefore used
as emulsifiers. They may also be contained in cosmetic preparations (e.g. shaving
foam).
Monoethanolamine is used as corrosion inhibitor in power stations as it maintains a
weakly alkaline pH and scavenges excess H+ ions.
150 Metrohm
Monograph
For their determination by ion chromatography the ethanolamines are protonated by the
addition of acid and can then be determined as their ammonium derivatives.
Learning content
• Determining inorganic and organic cations
• Changing the selectivity by adding complexing agents
• Determination of amines by ion chromatography
Eluent preparation
a) Dissolve 600 mg tartaric acid and 125 mg dipicolinic acid in 100 mL ultrapure water
under heating, then make up to 1 L with ultrapure water.
b) Dissolve 600 mg tartaric acid and 84 mg dipicolinic acid in 100 mL ultrapure water,
then make up to 1 L with ultrapure water.
152 Metrohm
Monograph
Chromatogram 63 Standard 2 – Eluent a
Remarks
If Eluent a is used then calcium and monoethanolamine coelute.
All solutions must be stored in plastic vessels. All contact with glass must be avoided for
a correct sodium determination.
154 Metrohm
Monograph
4.4.5 Experiment 22 – Alkali metals and alkaline earth metals in wine
The mineral substances in the grape juice are chiefly phosphates of calcium, magnesium and potassium.
The concentrations of these compounds are 3 – 5 g/L. The mineral content, also known as «ash» of the
wine is lower than that of the juice as a part of the minerals is taken up by the yeasts in their metabolism.
The mineral content is also reduced by the separation of tartrates. This means that wine contains only
about 1.8 – 2.5 g/L «ash». Important cationic constituents are given as oxides; these are K 2O (approx.
40%), MgO (approx. 6%), CaO (approx. 4%) and Na 2O (approx. 2%).
(see Experiment 12)
Learning content
• Analysis of foodstuffs
• Process analysis – checking tartrate precipitation
Eluent preparation:
Dissolve 600 mg tartaric acid and 125 mg dipicolinic acid in 100 mL ultrapure water
under heating, then make up to 1 L with ultrapure water.
Sample preparation:
Filter and dilute the sample (1:10).
Remarks
All solutions must be stored in plastic vessels. All contact with glass must be avoided for a correct
sodium determination. The pH of the standard and sample solutions must be between 2.5 and 3.5.
156 Metrohm
Monograph
Chromatogram 66 Determination of cations in white wine (1:10 dilution)
Remarks
Peaks No. 2, 3, 4 and 7 not evalued.
Remarks
Peaks No. 2, 3, 4, 5, 7 and 9 not evaluated.
158 Metrohm
Monograph
5 Literature cited
[1] H. Small, T.S. Stevens, W.C. Bauman, Anal. Chem. 47 (1975) 1801
[2] J. Weiß «Ionenchromatographie», 2. Aufl. (1991), Verlag Chemie, Weinheim
[3] J.M. Mermet, M. Otto, H.M. Widmer «Analytical Chemistry», 1st ed. (1998), Wiley-
VCH Verlag, Weinheim-New York
[4] P.R. Haddad, P.E. Jackson, «Ion Chromatography: Principles and Applications», 1 st
ed. (1990), J. Chromatogr. Library Vol. 46, Elsevier Verlag, Amsterdam
[5] J.S. Fritz, D.T. Gjerde, «Ion Chromatography», 3rd ed., Wiley-VCH, Weinheim, 2000
[6] G. Schwedt, «Chromatographische Methoden in der Anorganischen Analytik»,
1. Aufl. (1980), Dr. Alfred Hüthig Verlag, Heidelberg
[7] International Union of Pure and Applied Chemistry (IUPAC), Pure & Appl. Chem. 65 (1993), 819
[33] D.T. Gjerde, J.S. Fritz, G. Schmuckler, J. Chromatogr. A 186 (1979) 509
160 Metrohm
Monograph