EMI 3rd PDF
EMI 3rd PDF
EMI 3rd PDF
Experiment 1
Materials required:
Galvanometer
Cell
Rheostat
Ammeter of desired range
Resistance wire
Key
Screw gauge
Lab Procedure:
The shunt resistance required to convert the galvanometer into ammeter of range I is
calculated using the formula,
1
Where, ρ- the resistivity of material of the wire
r- the radius of the wire, which can be measured using a screw gauge.
The galvanometer with shunt resistance is connected in series to a battery through an ammeter, key
and rheostat.
Insert the key.
Adjust the rheostat and set the current reading I of the given ammeter at a particular value.
The reading of the galvanometer Ig’ is noted. Now, the current through the converted ammeter is
calculated using the relation,
2
Observations:
Shunt resistance,
Calculations:
Shunt resistance,
Experiment 2
Objective:
Apparatus Required:
2. Potentiometer
3. Sliding jockey
4. Mains cord
5. Patch cords
Theory:
A potentiometer instrument for measuring the potential (or voltage) in a circuit taps off a fraction
of a known voltage from a resistive slide wire and compares it with the unknown voltage by
means of a galvanometer. The potentiometer method is the usual basis for the calibration of
voltmeters, ammeters, and wattmeters. Since the potentiometer is a DC measurement device, the
instrument to be calibrated must be of the DC or electrodynamometer type. One of the first
requirements in this calibration procedure is that a suitable, stable DC supply be available, since
any variation in the supply voltage causes a corresponding change in the voltmeter calibration
voltage.
4
Diagram of Callibration of Voltmeter:
Procedure:
1. Connect the mains cord to the Trainer kit and switch On Mains Supply.
3. Note the output of standard DC supply (Vdc) by connecting terminal 32 to digital voltmeter
V1’s positive terminal and ground terminal 6 to negative of V1.
4. Once voltage is noted from V1, disconnect them and connect the negative terminal of
galvanometer G1 to positive terminal 32 of DC supply.
9. Touch jokey to X and then to Z terminals of potentiometer and see the reading of
galvanometer. Compare both reading of galvanometer.
5
10. Now slide the jokey on potentiometer wire and the find null point i.e., the point where
galvanometer G1 shows zero reading.
12. Set the voltage in analog DC Voltmeter (V) to some value (say 1 V) with the help of VR1
knob.
13. Touch jokey to X and then to Z terminals of potentiometer and see the reading of
galvanometer. Compare both reading of galvanometer.
14. Now slide the jokey on potentiometer wire and the find null point.
15. Now measure distance D (in cm) moved from terminal Z to null point.
Observation Table:
6
For Ammeter Calibration
L = [(n-1)*100 + r] cm.
n= number of wire from the Z terminal, for odd line of wire take reading from lower scale and
for even line wire take reading from upper scale.
L= Distance
Precaution:
7
Government Polytechnic, Muzaffarpur
ELECTRONIC MEASUREMENTANDINSTRUMENTATION
LAB
Experiment 3
Theory:-
The hays bridge is the modification of the Maxwell Bridge. This bridge uses a resistance in
series with the standard capacitor. The bridge has four resistive arms in which the arms one is
consists of the resister R1, Lx .The arm 2 is consists of the variable resistance R3.The low value
of the resistance is obtain by the low resistive arms of the bridge. The value of R4 and C4 is the
standard value of the capacitor and resistance. By using the unknown inductance having a
resistanceR1. R2, R3,R4-is the known non-inductive resistance and C4 is standard value of the
capacitor. The unknown value of inductance and Quality factor of the Bridge is obtained by
formula.
Basic AC bridges consist of four arms, source excitation and a balanced detector. Commonly
used detectors for AC bridges are:
(1)Head phones
(2)Vibration galvanometers
(3)Tunable amplifier detectors Vibration galvanometer is extremely useful at power and low
audio frequency ranges. Vibration galvanometers are manufactured to work at various frequency
ranging from 5 KHZ to 1 KHZ. But one most commonly used between 200HZ
Advantage-1) This Bridge gives very simple expression for unknown for High Q coil.
8
Disadvantage-1)The hays bridge is suited for the measurement of the High Q inductor.
2)It is used to find the inductor having the q value of the smaller then 10.
Procedure:
2)Connect unknown inductance LX1 in the circuit. Make all connections to complete the bridge.
Circuit Diagram:-
Precautions :
9
Government Polytechnic, Muzaffarpur
Experiment :3A
Transformer 230/15v
Bread board
Resistors
Variable Resistor
Inductors
Digital Multimeter
Theory:
This bridge circuit measures an inductance by comparison with a variable standard selfinductance.
10
At balance, L1 = R3L2/R4 , R1= R3(R2+r2)/R4.
Procedure:
5. If the selection of R2 is correct the balance point can be obtained at minimum position.
Observation:
11
Result: Actual and practical values of Inductances are found to be nearly equal.
12
Government Polytechnic, Muzaffarpur
Experiment :3B
Objective:
Apparatus:
Transformer 230/15v
Bread board
Resistors
Variable Resistor
Inductors
Capacitor
Digital Multimeter
Theory:
This bridge circuit measures an inductance by comparison with a standard variable capacitance . The
connections and the phasor diagrams for balance conditions are shown below.
13
Let, L1 = unknown inductance of resistance R1,
(R1+jωL1)(R4/1+jωC4R4) = R2R3
Procedure:
4. If the selection of R2 is correct, the balance point is observed at minimum voltage, for a
particular R1 and then increases by varying R3 in the same clockwise direction.
6. Observe the balance of bridge either using loud speaker or digital multimeter.
7. Connect the digital multimeter at the output of the detector. Alternately, adjust R3 and proper
selection of R2.
14
Sl No. R2 R3 C1 L1=R3L2/R4 True Value of
L1
15
Re
16
Result: Actual and practical values of Inductances are found to be nearly equal.
17
Government Polytechnic, Muzaffarpur
Experiment :3C
Objective:
Apparatus:
Transformer 230/15v
Bread board
Resistors
Variable Resistor
Capacitors
Digital Multimeter
Theory:
The bridge is the simplest of comparing two capacitances. The connections and the phasor
diagram of this bridge are shown below. Let
R4 = Non-inductive resistors.
The balance can be obtained by varying either R3 or R4. Resistors R1 and R2 are connected in
series with C1 and C2 respectively. r1 and r2 are small resistances representing the loss
component of the two capacitors.
18
At balance, (R1+ r1+ 1/jωC1) R4 = (R2+ r2+1/jωC2) R3
From which we have C1/C2 = R4/R3 . Figure b shows the phasor diagram of the bridge under
balance conditions. The angles δ1 and δ2 are the phase angles of capacitors C1and C2
respectively.
Dissipation factor for the capacitors are D1 = tan δ1 =ω C1r1 and D2 = tan δ2 =ω C2R2
D2 – D1 = ω C2(R1R4/R3 – R2)
Therefore, if the dissipation factor of one of the capacitors is known, the dissipation factor for the
other can be determined.
Procedure:
6. If the selection of R3 is correct the balance point can be obtained at minimum position.
19
8. Since, the unknown capacitance whose resistive effect would be made for capacitive form and
R2 is adjusted for minimum output.
Observation:
20
Government Polytechnic, Muzaffarpur
Experiment :4
Theory:
Between the stators, there is laminated rotor having slots and winding which supplies voltage to
the slide-wire circuit of the potentiometer. When current start flowing from stators, the rotating
field is developed around the rotor and due to it e.m.f. is induced in the rotor winding. The phase
21
displacement of the rotor emf is equal to rotor movement angle from its original position and it is
related to stator supply voltage. The whole arrangement of winding are done in such a way that
the magnitude of the induced emf in the rotor may change but it does not affect the phase angle
and it can be read on the scale fixed on the top of the instrument.
Therefore, resultant induced emf in the rotor winding due to two stator winding
22
Coordinate type Potentiometer:
In coordinate AC potentiometer, two separate potentiometers are caged in one circuit as shown
in the figure. The first one is named as the in-phase potentiometer which is used to measure the
in-phase factor of an unknown e.m.f. and the other one is named as quadrature potentiometer
which measures quadrature part of the unknown e.m.f. the sliding contact AA’ in the in-phase
potentiometer and BB’ in quadrature potentiometer are used for obtaining the desired current in
the circuit. By adjusting rheostat R and R’ and sliding contacts, the current in the quadrature
potentiometer becomes equal to the current in the in-phase potentiometer and a variable
galvanometer shows the null value. S1 and S2 are signs changing switches which are used to
change the polarity of the test voltage if it is required for balancing the potentiometer. There are
two step-down transformers T1 and T2 which isolate potentiometer from the line and give an
earthed screens protection between the winding. It also supplies 6 volts to potentiometers. Now
to measure unknown e.m.f. its terminals are connected across sliding contacts AA’ using selector
switch S3. By doing some adjustments in sliding contacts and rheostat, the whole circuit gets
balanced and galvanometer reads zero at the balanced condition. Now the in-phase component
VA of the unknown e.m.f. is obtained from the in-phase potentiometer and quadrature component
VB is obtained from quadrature potentiometer.
23
And the phase angle is given by
Applications of AC Potentiometer:
1. Measurement of self-inductance.
2. Calibration of voltmeter.
3. Calibration of Ammeter.
4. Calibration of watt meter.
24
Types & Construction:
Audio frequency chokes (AFC) – designed to block audio and power line frequencies while
allowing DC to pass
Radio frequency chokes (RFC) – designed to block radio frequencies while allowing audio
and DC to pass.
Audio frequency chokes:
Audio frequency chokes (AFC) usually have ferromagnetic cores to increase their inductance.
They are often constructed similarly to transformers, with laminated iron cores and an air gap. A
major use in the past was in power rectifiers and direct current motor controllers to produce
direct current (DC), where they were used in conjunction with large electrolytic capacitors to
remove the voltage ripple (AC) at the output DC. A rectifier circuit designed for a choke-output
filter may produce too much DC output voltage and subject the rectifier and filter capacitors to
excessive in-rush and ripple currents if the inductor is removed. However, modern electrolytic
capacitors with high ripple current ratings, and voltage regulators that remove more power
supply ripple than chokes could, have eliminated heavy, bulky chokes from mains frequency
power supplies. Smaller chokes are used in switching power supplies to remove the higher-
frequency switching transients from the output and sometimes from feeding back into the mains
input. They often have toroidal ferrite cores.
Some DIY car audio hobbyists use choke coils with car audio systems (specifically in the wiring
for a subwoofer, to remove high frequencies from the amplified signal).
Radio frequency chokes:
Radio frequency chokes (RFC) often have iron powder or ferrite cores. They are often wound in
complex patterns (basket winding) to reduce self-capacitance and proximity effect losses.
Chokes for even higher frequencies have non-magnetic cores and low inductance.
A modern form of choke used for eliminating digital RF noise from lines is the ferrite bead, a
cylindrical or torus-shaped core of ferrite slipped over a wire. These are often seen on computer
cables. A typical RF choke value could be 2 millihenries.
An MF or HF radio choke for tenths of an ampere, and a ferrite bead VHFchoke for several
amperes
25
A ferrite "bead" choke, consisting of a cylinder of ferrite encircling a computer power cord to
block electronic noise.
The common-mode choke, where two coils are wound on a single core, is useful for suppression
of electromagnetic interference (EMI) and radio frequency interference (RFI) from power supply lines and for
prevention of malfunctioning of power electronics device. It passes differential currents (equal but opposite),
while blocking common-mode currents.[2] The magnetic flux produced by differential-mode (DM) currents in the
core tend to cancel each other out since the windings are negative coupled. Thus, the choke presents little
inductance or impedance to DM currents. Normally this also means that the core will not saturate for large DM
currents and the maximum current rating is instead determined by the heating effect of the winding resistance.
The CM currents, however, see a high impedance due to the combined inductance of the positive coupled
windings.
A typical common-mode choke configuration. The common mode currents, I1 and I2, flowing in the
same direction through each of the choke windings, creates equal and in-phase magnetic fields
which add together. This results in the choke presenting a high impedance to the common mode
signal
When the CM choke is conducting CM current, most of the magnetic flux generated by the
windings is confined with the inductor core due to its high permeability. In this case, the
leakage flux, which is also the near magnetic field emission of the CM choke is low.
However, the DM current flowing through the windings will generate high emitted near
magnetic field since the windings are negative coupled in this case. To reduce the near
magnetic field emission, a twisted winding structure can be applied to the CM choke.
26
A balanced twisted windings CM choke
27
The equivalent current loops and the magnetic fields generated
We need to conduct a current to a certain inductor. And then, use a probe to measure the
near field emission. First of all, a signal generator is connected to an amplifier, serving as a
voltage source. The output of the amplifier is then connected to the measured inductor. To
monitor and control the current flowing through the inductor, a current clamp is used to
clamp the conducting wire. An oscilloscope is connected to the current clamp to show the
current waveform. A probe is then used to measure the flux in the air. A spectrum analyzer
is connected to it to collect the data
28
Government Polytechnic, Muzaffarpur
Experiment :5
Aim: To observe the loading effect of a multi-meter while measuring voltage across a low
resistance and high resistance.
Equipment/Components Required:
Theory:
When we use a voltmeter to measure the voltage in a circuit, we always assume its input
impedance is very large that it does not load the circuit and it will always indicate the correct
voltage in a circuit. Unfortunately this is not always the case.
29
In order for any instrument to provide a measurement, it must take a small amount of energy
from the circuit under test, and use this energy to obtain a reading. While the amount of
energy taken from most circuits is virtually undetectable, this is not always so. Consider the
circuit as shown in Figure 1. Ideally, the internal resistance of a voltmeter is infinitely large,
resulting in no circuit current. The voltage appearing across the voltmeter will be 10 V
which means that the voltmeter provides a correct reading of 10.0 V.
However, all voltmeters have some internal resistance. If the internal resistance of the
voltmeter was equal to the series resistance, 5.6 MW, then the voltage appearing across the
voltmeter would be half of the supply voltage, resulting in a reading of 5.0 V. If the internal
resistance of the voltmeter is even smaller, the voltmeter’s reading will be smaller too. The
degree to which a meter loads a circuit under test is called the loading effect and is
determined mathematically as:
In order to know whether the internal resistance of the meters is comparable with the circuit
under test, you must know the meter’s internal resistance. The internal resistance of the
DMM and VOM for its ammeter operation can be found in their operation manual.
PROCEDURES:
1.1 Refer to the manual of the DMM, obtain the internal resistance of the voltmeter and
record this value.
Rint = ______________Ω
1.2 The internal resistance of an analog meter is generally dependent on the voltage range
used. In order to determine the internal resistance, the manufacturer provides a specification
called the sensitivity, S which has units of kW/V. The internal resistance is then determined as
the product of sensitivity and the voltage range of the meter as:
Refer to the manual of the VOM, obtain the sensitivity of the meter. Determine the
correct range that we would use to measure a voltage of 10 V. Calculate and record
the internal resistance of meter on this range.
Vrange
S
Rint
30
1.3 Connect the circuit of Figure 3. Place the DMM voltmeter directly across the terminals
of the voltage source and adjust the voltage source for exactly 10 V which is the
unloaded voltage between terminals A and B.
1.4 Remove the DMM and connect it between terminals A and B. Measure the voltage
appearing between these terminals.
1.6 Replace the DMM with a VOM. Set the Vrange of the VOM to appropriate range to
measure 10 V. Measure the voltage appearing between these terminals.
1.8 Based on the test circuit, calculate the internal resistance of the VOM.
____________________________________________________________
____________________________________________________________
____________________________________________________________
31
1.9 Refer to Figure 4. Calculate the voltage across each of the resistors using the voltage
divider rule.
V1 (unloaded)
V2 (loaded)
1.10 Connect the circuit of Figure 4 and measure the voltages V1 and V2 with both the
DMM and VOM voltmeters.
Reading V1 V2
DMM
VOM
1.11 Calculate the corresponding loading effect of the DMM and the VOM
1.12 Replace the resistors 5.6 kW and 3.9 kW with 5.6 MW and 3.9 MW in Figure 4.
Calculate the voltage across each of the resistors and measure the voltage across each of
the resistors with DMM and VOM.
32
1.13 Calculate the corresponding loading effect of the DMM and the VOM.
33
Connect the circuit of Figure 5. Use the DMM as ammeter first and record the meter’s
reading. Repeat with a VOM ammeter. (Ensure that the ammeter is on the correct range
and polarity to measure the current).
2.5 Replace the 4.7 W resistor with a 47 W resistor and adjust the voltage of the DC supply to
3 V. Measure the current of the circuit with the DMM and VOM ammeters and
calculate the corresponding loading effect.
34
Government Polytechnic, Muzaffarpur
Experiment :6
Aim:Measurement of voltage, frequency, time period and phase angle using Cathode Ray
Oscilloscope (CRO).
Procedure:
Phase Measurement using Lissajous Patterns (X-Y Mode):
To Measure the phase difference of two sine waves their frequencies must be equal.
1. Connect a 1Volt peak-peak, 1KHz sine wave signal from the function generator to the
horizontal input of the CRO.
2. Connect the output of phase shift network to the vertical input as shown in figure.
3. Adjust the vertical and horizontal gains properly for good display.
4. Observe Lissajous Patterns for different combinations of R and C values.
36
! = tan-1 (f1/f2)
LISSAJOUS’ FIGURES
37
38
Government Polytechnic, Muzaffarpur
Experiment :7
Theory:
There are many methods for measurements of frequency or time. In our experiment only a few of
them are used: analog methods based on measurement of time with the oscilloscope, and direct
method based on of measurement frequency and time with the multifunction digital counter.
Oscilloscope method used in the experiments are extremely simple - they implement either
internal (linear) time base, or external reference signal (Lissajous method). Having known the
time base speed time/div (a value which may be read from oscilloscope’s screen), all we need to
do is to measure the length of one or more cycles of the observed signal.. This method is fast but
not very precise.
An other analog frequency measurement method involving the oscilloscope rely on using the
oscilloscope as a kind of null indicator for comparison of a sine signal with unknown frequency
with a reference sine signal whose frequency should be well defined and easily varied. One of
the signals is fed to the Y channel of the oscilloscope, the second one to the X channel. An
interaction of these two signals produces on the display more or less complicated snaky loops,
whose form allows for determining the unknown frequency. A typical Lissajous figure is shown
in Fig. 1
39
Fig. 1 Frequency measurement with the Lissajous method a) measurement diagram, b) Lissajous
figure and frequency determination.
If we assume that fy is the unknown frequency signal connected to the Y channel and fx is the
reference signal connected to the X channel then we have
where nx, ny denote number of intersections of the Lissajous curve with horizontal and vertical
axes, respectively, and dφ/dt is the rate of phase change (the trace rotation speed). The reference
signal should be adjusted until the displayed figure is possibly stable (dφ/dt ≈ 0). It is sometimes
attainable with difficulty, and needs both signal sources involved having adequate frequency
stability. The method's error is roughly equal to relative calibration error of the reference source.
Due to the mentioned stability problems, usefulness of the method is limited to rather low
frequency applications.
Digital counter methods rely on continuing number of events (in this case – the number of
cycles) with the counter open during precisely determined window. The periodic input signal of
any shape, including sine waveform, is formed in an input shaper block to have standard form of
possibly short pulses that are fed to the counter controlled by an accurate and stable quartz
oscillator (see Fig.2).
40
If the counter is opened for e.g. 1 second, then the number of pulses counted during this time
directly gives the measured frequency. If we denote open gate time as τ, input signal period as
Tx, and number of cycles counted as N, then
It is intuitively obvious that the accuracy of this method mainly depends on the accuracy of gate
timing. It may be shown that the limiting error of direct frequency measurement method is equal
to
where δgfref is the limiting error of quartz oscillator frequency.5gfk is the limiting.
In conclusion, one may see that the limiting error of direct frequency measurement method decreases
41
Fig. 1. Module F01. G1 – sinusoidal waveform output, G2 – rectangular waveform output, PF –
phase shifter
42
Note: Before the measurements:
- connect power supply to the F01 Module.
- set a frequency counter impedance to 50 Ω for inputs 1 and 2:
1.1 Use the DFC 53220A (Digital Frequency Counter) to measure time parameters (period and
frequency) of sine signal from generator G1 output in F01 module
a) PERIOD – measure the period of the signal with gating time of 1 s. If selected gating
time is longer than measured period, the DFC measures an average period.
b) FREQ – measure the frequency of the signal with gating time of 1 s. In this case, the
frequency is measured with the indirect method.
c) TOTALIZE:GATED – counting the number of pulses during 1 s of Gate Time. This case
realizes the direct frequency measurement method (using frequency definition – counting of
phenomena occurrences during reference time interval).
Compare and comment obtained results from PERIOD, FREQ and TOTALIZE:GATED
measurement method
1.2 Use the oscilloscope to measure time parameters (period and frequency) of sine signal from
generator G1 output in F01 module.
Take measurements of the period and frequency with the use the manual procedure (length
measurements) and the automatic measurement functions. Set the oscilloscope to minimize
measurements errors.
Insert the oscillograms into the report. Write down Cx constant and measurement results.
Estimate the limiting error of both measurements.
43
Compare and comment obtained results of period and frequency measurements with the
corresponding values obtained in point 1.1a and 1.1b. Assume the DFC as a reference
instrument.
2.1 Use the DFC 53220A (Digital Frequency Counter) to measure time parameters (period and
frequency) of sine signal from generator G2 output in F01 module.
Prior to the measurements assess the stability of the G2 generator and adjust the resolution of the
frequency counter. To do this, set the number of digits displayed on the frequency counter
display in a way to achieve observed instability of the result on the least significant digit
.
a) Measure the period and frequency using the PERIOD and FREQ functions respectively. Set
the gating time of 1 s.
b) Measure the pulse width of rectangular waveform (Width: Pos) and time interval between
pulses (Width: Neg).
c) Determine duty cycle of the signal based on the parameters obtained in point a) and b).
d) Measure the duty cycle of the signal using automatic function in the DFC (DutyCycle: Pos).
2.2 Use the oscilloscope to measure time parameters (period and frequency) of sine signal from
generator G2 output in F01 module
a) Take measurements of the period and frequency with the use the manual procedure (length
measurements) and the automatic measurement functions. Set the oscilloscope to minimize
measurements errors.
Insert the oscillograms into the report. Write down Cx constant and measurement results.
Estimate the limiting error of both measurements
Compare obtained results of period and frequency measurements with the corresponding values
obtained in point 2.1a. Assume the DFC as a reference instrument.
point 3.1.
Compare obtained results of period and frequency measurements with the corresponding values
obtained in point 2.1a. Assume the DFC as a reference instrument.
b) Take measurements of the pulse width and time interval with the use the manual procedure
and the automatic measurement functions (WIDTH+ and WIDTH-).
44
Insert the oscillograms into the report. Estimate the limiting error of the measurements.
c) Based on results in point a) and b), determine value of the duty cycle and estimate the limiting
error of both results.
d) Take measurements of the duty cycle with the use of automatic function +DUTY.
Compare values of duty cycle obtained in points 2.1c and 2.1d. Assume the DFC as a reference
instrument
3.1 Connect the output of the G1 generator to the input of the phase shifter PF in F01 module.
Use the DFC to measure the phase shift between signals on the input and output of the PF
module.
3.2 Use the oscilloscope to measure phase shift between signals on the input and output of the PF
module. Use the sine signal from G1 as an input signal for the phase shifter (PS).
Take measurements with the use the manual procedure (length measurements. Set the
oscilloscope to minimize measurements errors. Insert the oscillogram into the report and mark all
sectors used for ε calculation.
Compare and comment obtained result of phase shift measurement with the corresponding value
of phase shift obtained in point 3.1.
Conclusion::
In conclusion, one may see that the limiting error of direct frequency measurement method
decreases
45
Government Polytechnic, Muzaffarpur
Experiment :8
Aim: Measurement of rise, fall and delay times using a Cathode Ray Oscilloscope
Theory:
Rise Time
In the digital world, rise time measurements are critical. Rise time may be a more appropriate
performance consideration when you expect to measure digital signals, such as pulses and steps.
our oscilloscope must have sufficient rise time to accurately capture the details of rapid
transitions. Rise time describes the useful frequency range of an oscilloscope.
To calculate the oscilloscope rise time required for our signal type, use the following equation:
This basis for oscilloscope rise time selection is similar to that for bandwidth. As in the case of
bandwidth, achieving this rule of thumb may not always be possible given the extreme speeds of
today’s signals. Always remember that an oscilloscope with faster rise time will more accurately
capture the critical details of fast transitions. In some applications, we may know only the rise
time of a signal. A constant allows we to relate the bandwidth and rise time of the oscilloscope,
using the equation:
46
Rise time is the amount of time a pulse takes to go from a low to high voltage. By convention,
the rise time is measured from 10% to 90% of the full voltage of the pulse. This eliminates any
irregularities at the pulse’s transition corners. Pulse width is the amount of time the pulse takes to
go from low to high and back to low again. By convention, the pulse width is measured at 50%
of full voltage. Figure 69 illustrates these measurement points.
For rise time and fall time measurements, the 10% and 90% amplitude points are used as starting
and ending reference points.
Procedure:
1. Apply a signal to the INPUT jack. Set the vertical MODE to the channel to be used. Use the
VOLTS/DIV and VARIABLE to adjust the waveform peak-to-peak height to five divisions.
2. Using the vertical POSITION control and the other controls, adjust the display sich that the
wavedoem is centered vertically in the display. Set the SWEEP TIME/DIV to as fast a setting as
possible consistent with observation of both the 10% and 90% points. Set the SWEEP VARIABLE
control to CAL position.
3. Use the horizontal POSITION control to adjust the 10% point to coincide with a vertical
graduation line and measure the distance in divisions between the 10% and 90% points on the
waveform. Multiply this by the SWEEP TIME/DIV and also by 1/10 if "X10MAG" mode was used.
. NOTE:
The graticule on the CRT includes the 0, 10, 90, and 100 % lines assuming that 5 divisions
correspond to 100 %. Use them as a reference for accurate measurements.
Risetime = Horizontal distance (div) X (SWEEP TIME/DIV setting) / "X10 MAG" value.
47
Example
Rise time and fall time can be measured by making use of the alternate step 3 as described
below as well.
4. Use the Horizontal POSITION control to set the 10% point to coincide with the center vertical
graduation line and measure the horizontal distance to the point of the intersection of the
waveform with the center horizontal line. Let this distance be D1. Next adjust the waveform
position such that the 90% point coincides with the vertical centerline and measure the
distance from that line to the intersection of the waveform with the horizontal centerline.
This distance is D2 and the total horizontal distance is then D1 plus D2 for use in the above
relationship in calculating the rise time or fall time.
Risetime = (D1 + D2) (div) X (SWEEP TIME/DIV setting) / "X10 MAG" value.
48
Example
For the example, the measured D1 is 1.6 divisions while D2 is 1.4 divisions. If SWEEP
TIME/DIV is 2 us/div we use the following relationship
Conclusion:
49
Government Polytechnic, Muzaffarpur
Experiment :9
Apparatus Required:
Theory:
LCR meters are measuring instruments that measure a physical property known as impedance.
Impedance, which is expressed using the quantifier Z, indicates resistance to the flow of an AC
current. It can be calculated from the current I flowing to the measurement target and the voltage
V across the target’s terminals. Since impedance is expressed as a vector on a complex plane,
LCR meters measure not only the ratio of current and voltage RMS values, but also the phase
difference between current and voltage waveforms.
50
Typical equations for LCR meters
The test fixture used when measuring a target has residual components and can be expressed
using an equivalent circuit such as that shown in the figure below. Consequently, the measured
value Zm is expressed using an equation that contains these residual components, as shown
below. To calculate the true value Zx, it is necessary to calculate the open residual component
and short residual component and then correct the measured value. These correction processes
51
are known as open correction and short correction, respectively, and LCR meters include
functionality for performing both.
Zm:
Measured value
Zs:
Yo:
Zx:
The measurement signal output from the LCR meter is voltage-divided between the output
impedance R and the measurement target Zx. Thus the set measurement signal level V is not
applied as-is to the measurement target Zx. LCR meters have three measurement signal modes.
The user sets the measurement signal level V in the figure. This value is the voltage when the
measurement terminals are in the open state.
52
The user sets the value Vx in the figure (the voltage between the measurement target Zx’s
terminals). This mode is used when measuring targets that exhibit voltage dependence, for
example MLCCs with a high dielectric constant.
The user sets the value I in the figure (the current that flows to the measurement target Zx). This
mode is used when measuring measurement targets that exhibit current dependence, for example
inductors with cores.
Obsevation:
Conclusion:
53