Optical Pyrometer
Optical Pyrometer
Optical Pyrometer
This optical pyrometer compares the brightness of image produced by temperature source
with that of reference temperature lamp. The current in the lamp is adjusted until the
brightness of the lamp is equal to the brightness of the image produced by the temperature
source. Since the intensity of light of any wave length depends on the temperature of the
radiating object, the current passing through the lamp becomes a measure of the
temperature of the temperature source when calibrated.
An eye piece at one end and an objective lens at the other end.
A power source (battery), rheostat and millivoltmeter (to measure current) connected to a
reference temperature bulb.
An absorption screen is placed in between the objective lens and reference temperature
lamp. The absorption screen is used to increase the range of the temperature which can be
measured by the instrument.
The red filter between the eye piece and the lamp allows only a narrow band of wavelength
of around 0.65mui
Operation of optical pyrometer:
When a temperature source is to be measured , the radiation from the source are focused
onto the filament of the reference temperature lamp using the objective lens. Now the eye
piece is adjusted so that the filament of the reference temperature lamp is in sharp focus
and the filament is seen super imposed on the image of the temperature source. Now the
observer starts controlling the lamp current and the filament will appear dark as in figure (a)
if the filament is cooler than the temperature source, the filament will appear bright as in
figure (b) if the filament is hotter than the temperature source, the filament will not be seen
as in figure (c) if the filament and temperature source are in the same temperature.
Hence the observer should control the lamp current until the filament and the temperature
source have the same brightness which will be noticed when the filament disappears as in
figure (c) in the superimposed image of the temperature source [ that is the brightness of
the lamp and the temperature source are same]. At the instance, the current flowing through
the lamp which is indicated by the millivoltmeter connected to the lamp becomes a measure
of the temperature of the temperature source when calibrated.
Source:
http://instrumentationandcontrollers.blogspot.in/2010/09/optical-pyrometer-disappearing-
filament.html