Mouse IL-8 (Interleukin 8) ELISA Kit: Instruction Manual
Mouse IL-8 (Interleukin 8) ELISA Kit: Instruction Manual
Mouse IL-8 (Interleukin 8) ELISA Kit: Instruction Manual
Kit Components
Item Specifications(48T/96T) Storage
Micro ELISA Plate(Dismountable) 8×6 or 8×12 4°C/-20°C
Lyophilized Standard 1 vial or 2 vial 4°C/-20°C
Sample / Standard dilution buffer 10ml/20ml 4°C
Biotin-detection antibody (Concentrated) 60ul/120ul 4°C
Antibody dilution buffer 5ml/10ml 4°C
HRP-Streptavidin Conjugate(SABC) 60ul/120ul 4°C(shading light)
SABC dilution buffer 5ml/10ml 4°C
TMB substrate 5ml/10ml 4°C(shading light)
Stop solution 5ml/10ml 4°C
Wash buffer (25X) 15ml/30ml 4°C
Plate Sealer 3/5pieces
Product Description 1 copy
Automated Washing
Aspirate all wells, then wash plate THREE times with350ulwash buffer. After the final wash,
invert plate, and clap the plate on absorbent filter papers or other absorbent material. It is
recommended that the washer be set for a soaking time of 1 minute.
Note: The standard solutions are best used within 2 hours. The standard solution should be at
4°C for up to12 hours. Or store at -20 °C for up to 48 hours. Avoid repeated freeze-thaw cycles.
3, Preparation of Biotin-detection Antibody working solution
prepare within 1 hour before the experiment.
1) Calculate the total volume of the working solution: 0.1 ml / well × quantity of wells. (Allow
0.1-0.2 ml more than the total volume)
2) Dilute the Biotin-detection antibody with Antibody dilution buffer at 1:100 and mix
thoroughly.(i.e. Add 1 μl of Biotin-detection antibody into 99 μl of Antibody dilution buffer.)
4, Preparation of HRP-Streptavidin Conjugate (SABC) working solution:
prepare within 30min before the experiment.
1) Calculate the total volume of the working solution: 0.1 ml / well × quantity of wells. (Allow
0.1-0.2 ml more than the total volume)
2) Dilute the SABC with SABC dilution buffer at 1:100 and mix thoroughly. (i.e. Add 1 μl of
SABC into 99 μl of SABC dilution buffer.)
Assay Procedure
Before adding to wells, equilibrate the SABC working solution and TMB substrate for at least 30
min at room temperature (37 °C). When diluting samples and reagents, they must be mixed
completely and evenly. It is recommend to plot a standard curve for each test.
1. Set standard, test sample and control (zero) wells on the pre-coated plate respectively, and
then, record their positions. It is recommend to measure each standard and sample in
duplicate. Wash plate 2 times before adding standard, sample and control (zero) wells!
2. Aliquot 0.1ml of1000pg/ml,500pg/ml, 250pg/ml, 125pg/ml, 62.5pg/ml, 31.25pg/ml,
15.625pg/ml, standard solutions into the standard wells.
3. Add 0.1 ml of Sample / Standard dilution buffer into the control (zero) well.
4. Add 0.1 ml of properly diluted sample (Mouse serum, plasma, tissue homogenates and
other biological fluids.) into test sample wells.
Summary
1. Wash plate 2 times before adding standard, sample and control (zero) wells!
2. Add 100μL standard or sample to each well for 90 minutes at 37°C
3. add 100μL Biotin-detection antibody working solution to each well for 60 minutes at 37°C
4. Aspirate and wash 3 times
5. Add 100μLSABCworking solution to each well. Incubate for 30 minutes at 37°C
Specificity
This assay has high sensitivity and excellent specificity for detection of IL-8 . No significant
cross-reactivity or interference between IL-8 and analogues was observed.
Recovery
Matrices listed below were spiked with certain level of IL-8 and the recovery rates were
calculated by comparing the measured value to the expected amount of IL-8 in samples.
Linearity
The linearity of the kit was assayed by testing samples spiked with appropriate concentration of
IL-8 and their serial dilutions. The results were demonstrated by the percentage of calculated
concentration to the expected.
Precision
Intra-assay Precision (Precision within an assay): 3 samples with low, middle and high level IL-8
were tested 20 times on one plate, respectively.
Inter-assay Precision (Precision between assays): 3 samples with low, middle and high level IL-8
were tested on 3 different plates, 8 replicates in each plate.
CV(%) = SD/meanX100
Intra-Assay: CV<8%
Inter-Assay: CV<10%
Stability
The stability of ELISA kit is determined by the loss rate of activity. The loss rate of this kit is less
than 10% within the expiration date under appropriate storage condition.
To minimize extra influence on the performance, operation procedures and lab conditions,
especially room temperature, air humidity, incubator temperature should be strictly controlled.
It is also strongly suggested that the whole assay is performed by the same operator from the
beginning to the end.