AQS 1 User Guide 2018 PDF
AQS 1 User Guide 2018 PDF
AQS 1 User Guide 2018 PDF
The training website contains a comprehensive document library of technical notes and service and
calibration templates to download and print. The training website has detailed descriptions of service
activities and calibration and installation examples and is updated regularly. Video tutorials are
featured heavily. Go to training.aeroqual.com. Contact technical@aeroqual.com to receive access.
Where possible, references to further material will be provided throughout this document
Contact the Aeroqual distributor in your local country for technical support. If local technical support is
not available contact the Aeroqual support team: technical@aeroqual.com.
TSP inlet
Particle pump
Embedded Computer
Modem (Optional)
DETECTABLE
RESOLUTION
LINEARITY %
HOUR
LIMIT (PPB)
MODULE EXPECTED
MODULE
GAS RANGE
SENSOR
ZERO
LOWER
FLOW PRECISION Zero / (PPB) MODULE
OF F.S
MODULE (PPB) (PPM)
(PPB)
TYPE
RATE SPAN % OF LIFETIME
SIZE
SPAN % OF
FS
READING
Note: O3 and NO2 Modules have no serviceable parts. The VOC analyzer module uses a replaceable 10.6
eV UV lamp, if the lamp needs replacing contact the Aeroqual distributor in your local country for
technical support.
Gas modules come in two different designs. These two designs are shown schematically in Figure 1-
2. A list of which gas modules use which design is given in Table 1-1.
• Direct design (Ozone): The gas is passed directly to the sensor continuously.
• Zero switching design (VOC and NO2): The module uses a solenoid to switch between a
zero path which removes the target gas and creates a “zero” or baseline measurement, and a
measurement path which measures the ambient air. The final measurement in ppm is an
arithmetic combination of the zero and sample measurement. This technique is referred to as
Automatic Baseline Correction.
The gas flow rate is controlled using a critical orifice shown in red in Figure 1-2. The flow rate is listed
in Table 1-1.
Figure 1-4 Key components labelled (Modules may vary from that shown)
Each gas module contains: a gas inlet port (white ring),a gas exhaust port (blue ring), two 12 V power
connectors to allow the module to be connected to a 12 V power bus, two RJ-45 connections to allow
the module to be connected to the RS-485 communications bus, and a module status LED.
The sample flow for the Particle Monitor and Particle Profiler is separate from the gas sample flow.
There is a separate pump, a separate inlet and a separate exhaust.
Figure 1-6 The Particle Monitor uses three modules: a) the optical engine on the far right, b) the pump
module in the middle, c) the I/O module on the left
The particulate size selection (TSP or PM10 or PM2.5 or PM1.0) is determined by the sharp cut cyclone
on the Particle Monitor inlet on the outside of the AQS 1 shown in Figure 1-7.The Particle Monitor can
be configured to measure either of these four size fractions simply by changing the inlet cyclone and
making a small change in the software.
Figure 1-8 The Particle Profiler uses three modules: a) the optical module on the far right, b) the pump
module in the middle, c) the electronics module on the left
• The Particle Profiler measures particulate concentrations using particle counting, it outputs
multiple size fractions at the same time.
• The Particle Profiler can measure TSP and PM10 and PM2.5 and PM1 without requiring a sharp
cut cyclone.
Factory calibration period Every two years (24 months) Every year (12 months)
MetOne MSO Cirrus MK427 Vaisala WXT536 Gill Instruments Li-Cor LI-200
weather station noise meter weather station Windsonic Pyranometer
Figure 1-9 The external sensors are connected to the green plug on the Aux module. The cable passes
through water proof glands on the bottom of the AQS 1
Details on how to mount each external sensor and how each sensor is wired to the AUX module are
given in the online training:
12 V DC fuse pushed in, the “on” position 12 V DC fuse pulled out, the “off” position
Figure 1-10 The fuse holder next to the 12 V rail can be pulled out to turn off the AQS 1 while mains
power is attached.
• Remove the protective film before pushing the fuse back in (turning the instrument on).
• After the PC has booted up, the AQS 1 will appear as a WIFI network in your PCs list of WIFI
networks. Log on to the WIFI network with password “Aeroqual”.
Figure 1-11 Remove the protective film from the PM before starting. Once you have started check your
WIFI network list
• Then open a browser window and type 10.10.0.1 in the address bar. Login to Connect using
login name: administrator password aqmadmin
NOTE: See Section 3.1: Connecting to your AQS 1 for the first time for more detail
Figure 1-12 Aeroqual Connect is the operating system for the AQS 1. Aeroqual Connect is accessed
using a web browser.
Figure 1-14 Select multiple instruments by region, and plot data from multiple different instruments on
the same graph.
Figure 1-15 Generate advanced graphical presentations such as wind roses and pollution roses.
Advanced support from your distributor and the Aeroqual technical support team
Aeroqual Cloud allows the technical support team to provide fast and comprehensive technical
support directly to the end user. Instrument diagnostics and operational parameters are immediately
available to Aeroqual engineers through Aeroqual Cloud.
• You need to supply a mains power cable and local plug, Aeroqual recommends a residual
current device (RCD) be used to protect against power spikes.
• You will only need a Phillips head screwdriver for assembly in the laboratory or office.
• There is no software to download and install, and no cables are require for communication.
Communication to the AQS 1 is through your web browser on your laptop, tablet or smart
phone using WIFI.
• If you have purchased a MOXA modem, this needs a local SIM (dynamic IP) same as you
would have in your smart phone, but you must know the APN of your chosen
telecommunications provider, you can look this up: www.apnchanger.org
2.1 Unpacking
The AQS 1 is packaged in a cardboard carton with moulded protective inserts.
Dimensions: 62 x 32 x 55 cm (L x W x H)
Weight: < 12.5 kg (Configuration used for weight calculations: base unit,
nephelometer, PM10 sharp cut, O3 module, modem, heater off / heater on.
Additional components such as weather or noise sensors will be packaged in
additional boxes.
a) Examine the ShockWatch® label on the side of the shipping box. If the
indicator is red do not refuse the shipment. Make a notification on
delivery receipt and inspect for damage. If damage is discovered, leave
item in original packaging and request immediate inspection from carrier
within 15 days of delivery date (3 days international).
b) Verify the serial number label on the documentation matches the serial
label on the AQS 1.
c) Verify that all components have been shipped as per the packing slip.
Contact your Distributor or Aeroqual if you suspect any parts are missing.
Caution: The high voltage mains supply must be wired by a certified electrician in
compliance with local electrical regulations.
The AQS internal power requires regulated 12 VDC to maintain a constant pump speed. An external
mains to 12VDC power supply (Meanwell HLG80-12A 60W 12V) is fitted on the outside of the
enclosure and comes with an IP rated outdoor electrical plug. You must wire the other end of the plug
to mains power.
Note: Aeroqual does not supply the mains power cable or the electrical plug for your local power
socket, you must supply this yourself.
1. Locate the power connector terminal which is situated outside the enclosure at the bottom.
The power supply will already be wired to the connector however the cable to connect to
mains power needs to be wired.
Nut gland
Plug
Neutral
Connector (Blue)
Terminal block
Live
Earth (Brown)
(Yellow/Green)
2. Feed the power cable through the nut gland 3. Reconnect into the connector
plugs and connector and wire the cable to the ensuring the plugs are fitted securely
terminal block according to the diagram. and the nut gland tightened.
Note: To remove the terminal block from inside the connector unclick it from the casing by
turning it slightly then pull out
Pull the 12 VDC fuse out from the fuse holder “off position” before testing the mains connection.
12 V DC fuse pushed in, the “on” position 12 V DC fuse pulled out, the “off” position
Figure 2-1 The 12V DC fuse can be pulled out to turn off the AQS 1 or pushed in to power the AQS 1
Once you have removed the PM power then push in the 12 DC fuse to the “on” position as shown in
12 V DC fuse pushed in, the “on” position 12 V DC fuse pulled out, the “off” position
• Turn off power to the AQS 1 during this process by pulling out the 12 DC fuse.
• The Particle Monitor ships from the factory with a thin film of tape which protects the inlet, this
must first be removed as shown in Figure 2-2.
Remove protective
tape
• The inlet is wrapped in a thin film heater which has a white plug on the end.
• Carefully thread the white plug through the PM inlet port on the roof of the AQS 1.
Figure 2-4 Gently feed in the power cable. The aluminium tube must push all the way in to the PM
module.
• It is very important to make sure that the aluminium inlet tube is pushed all the way into the
optical engine block.
• If the aluminium tube does not fully push down in to the optical engine block, then create
more length in the aluminium tube by removing the grey retaining nut and then sliding up the
rubber grommet on the top of the inlet. Replace the nut after you have done this.
• You need to make sure you have correctly and securely inserted the aluminium tube fully in to
the optical engine block.
• Push the fuse holder in to the “on” position and connect the white power plug to the module
to begin flowing air through the inlet.
• Then perform a leak check and flow check. See Section 4.4.5 and 4.4.6.
Figure 2-6 Apply power to the PM module, then fit the vacuum gauge to the inlet to test for leaks.
• When you are satisfied the inlet is fully pushed in to the engine block, then connect the heater
plug on the inlet to the plug coming from the module.
• Insert the three supplied 6/32 UNC retaining screws to secure the inlet to the aluminium
retaining ring
Figure 2-7 During field installation, secure the PM inlet using the retaining screws and plug in the heater
• Attach the inlet and perform a flow check to make sure the inlet is fitted correctly and the
sample pump is working properly.
• When the AQS 1 is sampling ambient air in the field, a Kynar inlet fitting with mesh is attached
to the Swagelok nut at the end of the sample cane.
NOTE: Detailed instructions on gas flow measurement are given in Section 4.4.3.
• External sensors such as weather stations and noise meters need to be wired to the green
connector plug on the front of the auxiliary module. To connect third party sensors such as
weather sensors, feed the cable through a free cable gland and wire to the green plug.
• Each third party sensor requires a different wiring configuration. See online reference.
NOTE: To check the operation of the third party sensor in your office or laboratory prior to site
installation it is not required to feed the cable through the gland.
• If you are happy with the wiring then apply power, and log on to the Aeroqual Connect
software as described in section 3 Connect Software. After some minutes you should see the
parameters associated with the third party sensor being shown in the software.
• Aeroqual supplies an optional cellular modem (MOXA G3111 HSPA) for remote
communication. This is Aeroqual part number R38. This must be correctly configured to work
on the local telecommunications network in your country.
• The modem requires a standard SIM card (dynamic IP) which you can buy from a cell phone
store. The SIM card fits in a slot underneath the modem as shown in Figure 2-7.
• You must know the APN of your cellular provider, you can find out the APN by going to
www.APNChanger.org and looking up your country and cellular provider.
• Then enter the APN in to the correct field through the MOXA configuration interface shown in
Figure 2-7.
Figure 2-8 A SIM card must be inserted in to the modem and the Modem must be configured
Figure 3-1 The AQS 1 is controlled by an ePC, a modem can be included under the PC, on the right is the
website which is hosted locally on the ePC.
Cloud software
In addition to Aeroqual Connect software, which is running locally on the AQS 1, Aeroqual also offers
cloud based software called Aeroqual Cloud.
Aeroqual Cloud allows the data from multiple AQS 1 to be visualised on one screen. Users can
receive email or text alerts from Aeroqual Cloud. The AQS 1 must be connected to Aeroqual Cloud in
order to receive software updates and bug fixes.
Aeroqual technical support requires that the instrument be connected to Aeroqual Cloud in order to
provide fast and effective customer support.
NOTE: If the AQS 1 is not connected to Aeroqual Cloud, technical support from Aeroqual may be
limited and may be charged for.
Figure 3-2 In WIFI access point mode the AQS 1 appears as a WIFI network in the lit of WIFI networks on
your PC.
The next step is to open a blank browser window and enter 10.10.0.1 in to the address bar.
Figure 3-3 The AQS 1 default access login, this should be removed after another admin account has been
created
Figure 3-4 Column three in the Configure instrument app control the communication and connection
settings
There are two independent connection types; WIFI and wired Ethernet.
1. The ePC has built in WIFI which can be configured either:
• as an access point or
• it can be used to access an existing LAN over WIFI
2. The ePC has a separate wired Ethernet connection, this can used in a number of ways:
• To directly connect to a PC
• To connect to a wired LAN via a router
• To connect to an IP modem to provide cellular communication.
The WIFI connection and Ethernet connection are independent from each other so that both can be
used at the same time.
The recommended configuration for connection is to set the WIFI mode to access point and the wired
Ethernet to Client mode to use with the MOXA modem.
WIFI Wired Ethernet
Direct Connection Used by engineers Used by engineers at
These are commonly used by an engineer at location location performing
who is on site with the AQS 1 to perform performing maintenance
maintenance or calibration. maintenance
(recommended)
Remote connection The AQS 1 WIFI is The AQS 1 wired Ethernet
These are used by data analysts who need connected to an is connected to an existing
to download data, but don’t need to visit the existing WIFI wired Ethernet network,
installation site. network which could also be a
cellular modem
Figure 3-5 To establish a direct connection, a laptop must be connected to the ePC using an Ethernet
cable
Figure 3-7 (c) shows the communications settings for connection using an Ethernet cable. Note that
the Ethernet mode is set to DCHP (server) in this method.
When this happens the AQS 1 will automatically be assigned an IP address by the router.
You must find out what this IP address is so you can connect to the AQS 1. The WIFI or wired LAN IP
address will likely be in the format: 192.168.XXX.XXX
Figure 3-4 shows the Connect software with the assigned IP address in the address bar in the
browser.
If your laptop also has access to the same LAN, then this IP address can be typed into a browser
address bar and a connection to the AQS 1 can be established over the LAN connection.
Figure 3-6 The MOXA modem uses a SIM card and it must be configured for use with the AQS 1
The Ethernet setting must be set to Auto DHCP client to use the MOXA modem, as shown in Figure
3-7 (c).
Figure 3-7 The different connection types require different settings in the Configure Instrument app
a)
The Ethernet mode is set to Auto (DHCP client). In this configuration the MOXA modem can be used.
The MOXA modem will assign an IP address to the ePC (192.168.127.10)
The WIFI mode is set to Access Point so a direct connection to the EPC can be made over local WIFI
c)
The Ethernet mode is set to Auto (DHCP client). In this configuration the MOXA modem can be used.
The MOXA modem will assign an IP address to the ePC (192.168.127.10)
The WIFI mode is set to Client and is connected the office WIFI network. The ePC has been given a
WIFI IP address by the router (192.168.0.158). This IP address can be entered directly in the browser
to connect to the AQS 1.
d)
The Ethernet mode is set to Direct (DHCP Server). In this configuration a laptop can be directly
connected to the ePC using an Ethernet cable. Enter 10.10.0.1 to connect directly to the AQS 1.
Then insert the USB drive into any of the USB ports on the ePC.
Wait several minutes, the ePC will reboot automatically and then you should see a WIFI network
appear called “Aeroqual AQM” with password Aeroqual.
You can then remove the USB stick.
Figure 3-9 The landing page of Connect after login shows the available apps
Figure 3-10 The charts tool, in the Manage Data app shows each data channel in a stack of charts
3.4.1.2 Table
The data from each module can be viewed on screen in a Table. The from and to dates can be
selected using a calendar tool, and the averaging period such as daily, hourly or minute can also be
selected.
Figure 3-11 The table tool in the Manage Data app shows the data in a table with user configured date
range and averaging
Figure 3-12 The download data tool in the Manage Data app allows the data to be downloaded at user
configured date range and averaging
NOTE: You might need to manual sync multiple times. The first time the From Date is 1900, then it
creates a file with a maximum number of records of 1 million. Therefore the file might only
synchronize for a limited dataset. At the end of that sync the From Date will show up to where
the last data was downloaded. You run the sync command again and it gives you a new file
starting at that date and including the max number of records. You keep running this process
until you get to today’s date.
These files are then uploaded one by one using the manual sync upload tool in Cloud.
Figure 3-13 The manual sync tool in the Manage Data app allow all data from the AQS 1 to be manually
uploaded to the Cloud if a local data connection is not available
Diagnostics
Each module can be selected and real time module operation parameters can be viewed.
Module settings
Each module has a set of operational parameters which control the module. These can be viewed or
edited using the module settings dialogue. These settings should not be changed without instruction
from Aeroqual.
Figure 3-14 The Diagnostics and Advanced app displays real time instrument operational readings and
settings which control how the modules are controlled.
NOTE: maintaining accurate and complete Journal entries is a pre-requisite of achieving Advanced
Engineer status, a certification provided by Aeroqual to distributor engineers. Keeping service
records is also a requirement under Aeroqual’s factory warranty
Figure 3-15 The Journal automatically logs software changes, it also allows manual user entries
IMPORTANT!
These access credentials are published in multiple public documents and are only used to first
connect to and configure the AQS 1. Aeroqual recommends that a new Administrator user is
created with secure password and the default Administrator user is deleted which will make the
AQS 1 more secure.
There are three different user permissions which can be applied to individual users on the AQS 1.
Administrator
• Can control access login and password for other users.
• Has full control over all instrument settings.
• Can view and download data.
An administrator can access all of the apps in Connect, and is the only user type who can access the
Administration app. New users are created using the Administration App.
Engineer
• Has full control over all instrument settings.
• Can view and download data.
An engineer cannot access the Administration App.
User
• Can view and download data.
• Can perform a calibration and change calibration settings
For some of the service activities described a video is also available on vimeo.com. If a video
is available, a web-link will be provided next to this symbol.
Aeroqual recommends either the Bios Defender DryCal or the Aeroqual R7 flow meter, these can both
be used to measure gas flow and Particle Monitor flow.
Figure 4-1 A high quality flow meter is required for gas flow measurement of the AQS 1. Aeroqual
recommends the Bios DryCal or the TSI 4140 (AQS R7)
Figure 4-2 The AQS 1 can be put into Service Mode by clicking the button in the Calibration and Service
app. This will label the data as "Service" in the inlet column so the data can be filtered.
4.3.2 Journal
All service activities should be documented. The Journal is accessed by clicking the Journal button
next to the username at the top right of the screen, then selecting user entry.
Figure 4-3 All service activities can be recorded in the Journal manually.
NOTE: Whenever these service activities are performed it is important to record the activity in the
Journal using a manual user entry. See Section 4.3.2
Do not turn off the AQS 1, this procedure can be done with the AQS 1 turned on.
NOTE: It is very important to check the AQS 1 inlet flow rate is correct after changing the filter to
ensure there are no leaks
Figure 4-4 The particle filter on the gas inlet can be easily replaced, it should be replaced when it gets
dirty as shown in the image on the right (a new filter is shown on the left, dirty filter on the right).
Frequency of
Recommended Monthly.
service
The AQS 1 inlet rate is controlled, See Table 1-1 for a description of the expected flow rate for the
module.
When the AQS 1 is new, the inlet flow rate is recorded in the instrument log book which is provided
from the Aeroqual factory.
The measured flow rate should be equal to the previously measured flow rate which is recorded in the
instrument journal.
Figure 4-5 To measure the flow rate, remove the inlet filter and attach the inlet flow adaptor.
Figure 4-6.
Figure 4-6 The flow adaptor valve is used to adjust the AQS 1 inlet flow to the correct value.
If the correct flow rate cannot be set then it may be required to replace the gas sample pump. The
pump is part number AQS R10. The pump can easily be removed from the bracket for replacement. It
is important to use fresh Tygon Tube (CS Tube 10) when re-connecting the tube back on the pump.
This is because the barbs on the pump stretch the tube.
The online training website contains details on how to replace the gas sample pump.
Parts required AQS R31 (Particle Monitor) AQS R32 Particle Profiler
The filters for the Particle Monitor are located inside the optical module. Use a large flat head screw
driver to unscrew the filter housing on the front panel of the optical module.
The filters for the Particle Profiler are located in a filter casing outside the optical module and within
the Particle Profiler Pump Module.
Figure 4-7 The Particle Monitor (shown on the left) requires AQS R31 and the Particle Profiler (shown on
the right) requires AQS R32
Figure 4-8 The AQS R31 sample filter should be replaced once the filter begins to blacken
4.4.5 Flow check and flow adjustment of the Particle Monitor and Particle
Profiler
It is important to maintain the correct flow rate through the Particle Monitor or Particle Profiler. The
flow rate of the Particle Monitor and Particle Profiler are different
NOTE: The flow rate of the Particle Monitor is 2.0 ± 0.05 LPM
The flow rate of the Particle Profiler is 1.0 ± 0.05 LPM
Tools required Small Philips head screw driver, AQS R7 AQS R56
Frequency of
Recommended Monthly. (See Section 4.6)
service
Online
training.aeroqual.com: Dust Monitor technical Training Section 4.1
reference
The AQS R56 includes a flow adaptor which is fitted on the inlet, for the Particle Monitor it is fitted
above the cyclone as shown in Figure 4-9.
4.4.5.1 Using the buffer chamber when checking the flow of the Particle Profiler
When measuring (or adjusting) the flow rate of the Particle Profiler it is important to use the buffer
chamber to ensure accurate measurement. The buffer chamber is part of the AQS R56 Zero Filter
and Flow assembly. It is also part of the R20 Service kit. Figure 4-10 shows the buffer chamber
installed between the flow meter and the profiler inlet.
Figure 4-10 When measuring the flow on the Particle Profiler it is important to use the flow buffer
chamber
Exhaust valve
Figure 4-11 The flow rate of the Particle Monitor is set by adjusting the purge and exhaust valves
The flow rate of the Particle Monitor must be 2.0 LPM ± 0.05 (between 1.95 and 2.05 LPM)
• Fully close the purge valve by pushing the valve handle towards the module
• Adjust the exhaust valve until the flow reads 2.2 LPM
• Adjust the purge valve until the flow reads 2.0 LPM ± 0.05
The final adjustment can be a little tricky, turn the purge valve slowly. If it is not possible to reach 2.2
LPM at step 2 then there may be a leak or the pump might need to be replaced.
Exhaust valve
Figure 4-12 The flow rate of the Particle Profiler is set by adjusting the flow adjustment valve
The flow rate of the Particle Profiler is adjusted to 1.0 LPM ± 0.05 LPM using the flow adjustment
valve as shown in Figure 4-12.
Frequency of
Recommended Every three months (See Section 4.6)
service
Online
training.aeroqual.com: Dust Monitor technical Training Section 4.2
reference
Figure 4-13 Fit the vacuum gauge, adjust the valves. Then stop the pump by removing power from the
module, observe the change in pressure. If the leak rate is greater than 10 kPa in 10 seconds then there
may be a leak in the system. Check all the tube connections.
NOTE: The pressure change (leak rate) should be no more than 10 kPa in 10 seconds.
Figure 4-14 Cap the purge and exhaust lines. Place the vacuum gauge on the inlet. Disconnect the power,
observe the change in pressure.
NOTE: The pressure change (leak rate) should be no more than 10 kPa in 10 seconds
Frequency of
Recommended Monthly. (See Section 4.6)
service
Check that the gain and offset values for the Particle Monitor are set to sensible settings. The gain
should be between about 0.6 and 4.0. The offset should be 0.000.
See Section 6 for a discussion on gain adjustment of the Particle Monitor.
Figure 4-15 The zero filter is fitted above the cyclone for a Particle Monitor (left) and directly on the inlet
for a Particle Profiler (right)
Figure 4-16 When measuring the sample flow of the Particle Monitor a vacuum is measured. When
measuring the zero flow a positive pressure is measured.
The auto zero can be executed manually by temporarily changing one of the module settings. Once
the auto zero cycle is complete the setting needs to be returned back to the original setting.
• Go to the Diagnostic and Advanced app in Connect software.
• Select Module settings from the list on the left column
• Locate the Particle Monitor settings, these will be either PM10.0 or PM2.5
• Locate the TIMA parameter (it should be 720).
• Change this value to 1 by clicking in the cell and typing 1, then save the setting change.
This will cause the auto zero to start. The flow will change from 2.000 LPM negative flow into the inlet
to approximately 0.2 to 0.6 positive flow out from the inlet.
You will need to change the port on the flow meter to measure this flow as shown in Figure 4-15.
• If no flow is measured there may be an issue with the zero cycle or zero pump. See
Appendix 1 for trouble shooting.
• Measure the positive flow coming out from the inlet and record this in the journal.
The zero cycle takes approximately 6 minutes to complete.
NOTE: Once the flow has been measured the TIMA parameter must be set back to 720
Frequency of
Recommended 6 Monthly. (See Section 4.6)
service
The TSP head, particle trap and cyclone disassemble easily for cleaning.
Figure 4-17 The TSP head and cyclone can be disassembled for cleaning.
Typically the pumps last between 12 and 18 months. However the pumps should be replaced when
the correct flow can no longer be maintained.
Figure 4-18 The Particle Monitor contains two pumps, the Particle Profiler has one pump.
Frequency of
Only as required. (See Section 4.6)
service
Figure 4-19 A gas module can be checked for leaks by using a diaphragm pump and R7 flow meter
NOTE: If the module is found to have a leak, contact the Aeroqual distributor in your local country for
technical support.
Tools required Tube Cutter, Large Phillips head screw driver, AQS R7
If a module is removed or added then a software change is required. The software module
configuration can be found in the configure instrument app. Sensor modules can be added or
removed using this tool, shown in Figure 4-20.
Figure 4-20 Change the module configuration in the software in the configure instrument app.
IMPORTANT!
Make sure the polarity of the red and black cable is the correct orientation.
• Cap off the vacuum and sample lines using a luer cap
• Remove the module from the sensor list in the software (Configure instrument app)
• Measure the total inlet flow rate and document this in the journal
Figure 4-21 Adding a new module requires adding additional sample and exhaust lines.
Figure 4-22 Filters should be changed before they get too dirty. The frequency of filter changes will
depend upon local site conditions.
Data quality objectives will also influence the service frequencies for some operations. In monitoring
situations where data quality objectives are high, for example where monitoring is performed for
compliance or where a high amount of scrutiny will be placed on the data, service frequencies may be
higher.
It is recommended that the instrument engineer, over time, develops an appropriate servicing
schedule for a particular site.
Table 4-1 provides guidelines for service frequencies for the service activities documented in this user
guide.
Advantages
The Advanced Engineer certification allows Aeroqual
to assure end users of Aeroqual products that high
quality service can be expected from the engineer and
distributor performing the service.
Certification is valid for one year and is reviewed annually by examining instrument service records in
the journal and by discussion between Aeroqual and the engineer seeking re-certification.
Certification is only renewed if the engineer continues to demonstrate on-going high quality service.
• They must process warranties and Return Material Authorisations (RMAs) correctly and
quickly for end users.
• They must maintain contact with Aeroqual and provide feedback from their users which
Aeroqual can use to improve products and services.
• Aeroqual Advanced Engineers can use the Aeroqual Cloud Journal on cloud.aeroqual.com to
provide evidence that they are continuing to provide good service to end users.
• The Cloud journal must be used to record all service and maintenance visits.
• The Cloud journal can be used even if the instrument is not connected to Cloud (offline).
Aeroqual will make regular checks of the instrument journal on Cloud to see if required
scheduled service visits are being performed.
2. Field calibration. The AQS 1 is designed so that the gas module can be calibrated in the AQS
1 in the field. The field calibration set up and process is very similar to calibration of US / EU
EPA approved ‘reference’ analyzers.
b) periodical validation of the data is required (typically for advanced users only).
3. Factory calibration post-shipment. At any time the gas module can be removed from the AQS
1 and returned to the factory or one of our authorised service providers for factory calibration.
If the gas module has not been field calibrated then it is recommended to have the gas
module factory calibrated every 12 months.
• Aeroqual provides all of the equipment required to perform field calibration. Aeroqual offers a
portable calibrator called the AirCal 1000.
• Offset and gain adjustments are performed using the software tools provided by Aeroqual
Connect or Aeroqual Cloud.
• If the AirCal 1000 is being used, there is a separate PC desktop software application which
can be downloaded from the website.
• For ozone, an ozone generator, (ozone calibrator) must be used, Aeroqual recommends the
2B Tech Ozone generator part number AQM O3CAL.
• Aeroqual does not recommend calibration when first commissioning a new AQS 1. We
recommend first calibrating 1-3 months after commissioning.
• The quality of the calibration gas makes a big difference to success of a calibration, only
purchase calibration gas from a reputable calibration gas supplier.
• The AQS 1 must be in a stable condition before calibration can be attempted. The AQS 1
must have been running for at least 6 hours before calibration, ideally 24 hours.
• Record keeping is a crucial part of calibration, the journal feature in Connect and Cloud is
where calibration record keeping is performed.
• There are calibration forms in Appendix 5 which can be used to record data from this
calibration.
Calibration is discussed in detail in the online training. Several supporting videos are also available.
https://vimeo.com/152928037 https://vimeo.com/152913736
Online video
https://vimeo.com/152956203
Figure 5-1 Centre: O3Cal Ozone calibrator. Left: AirCal 1000 with NO2 cylinder from Calgaz, Right AQS 1,
Note: Both the AirCal 1000 and the Ozone calibrator take time to warm up and become stable.
• Turn on the AirCal 1000 and switch on the pump override switch on the back panel for at
least 30 mins before staring the calibration.
• Turn on the Ozone generator and set it to deliver 0.1 ppm Ozone for at least 30 mins
before starting the calibration, this allows time for the lamp in the ozone generator to
stabilise at 0.1 ppm.
See Table 5-1 for recommended cylinders (part numbers) from Calgaz.
A two-stage pressure regulator is required for use with these gas cylinders. Aeroqual can supply the
regulator, part number AIC GASREG01, for use with C10 fitting. The Aeroqual regulator is sown in
Figure 5-1, attached to the Calgaz cylinder.
Figure 5-2 High quality flow meters. Left: Bios DryCal, Right: Aeroqual R7 flowmeter.
Aeroqual makes the following recommendations about calibration frequency for the AQS 1:
• Follow all local regulatory calibration requirements if these are defined for the monitoring activity.
• Plan routine calibrations at an interval that is consistent with the data quality objectives for your
monitoring purpose.
• Field calibration may be necessary if the gas module is operating out of specification and all other
trouble-shooting steps have been taken.
The Journal feature in Connect/Cloud can also be used to record observations and make comments
throughout the calibration process.
Assumptions
This SOP assumes the instrument and calibration equipment are in good working condition and the
instrument is “ready for calibration”.
• The instrument has been running for at least 6 hours.
• The gas lines and inlet filter is clean.
• You are familiar with the operation of all of the calibration equipment.
• All calibration equipment, such as calibrators and gas bottles are working and within
calibration where necessary.
Note: Both the AirCal 1000 and the Ozone calibrator take time to warm up and become stable.
• Turn on the AirCal 1000 and switch on the pump override switch on the back panel for at
least 30 mins before staring the calibration.
• Turn on the Ozone generator and set it to deliver 0.1 ppm Ozone for at least 30 mins
before starting the calibration, this allow time for the lamp in the ozone generator to
stabilise at 0.1 ppm.
Figure 5-3 The calibration screen is where the average and standard deviation can be read. This is also
where the gain and offset can be read and set.
Figure 5-4 The a value on the NO2 module is set in the Manual Entry section of the Calibration and
Service App
Figure 5-3 shows the main calibration screen. This is where you will find the Average and Std Dev
readings which are used during the calibration. This is also the screen where you can see the current
gain and current offset values, and where you will change the gain and offset values, if required,
during the calibration.
Figure 5-4 is the manual entry screen, where you can also change gain and offset and the a value of
the NO2 module.
• Turn on the AirCal 1000 and switch on the pump override switch on the back panel for at least 30
mins before staring the calibration.
• Turn on the Ozone generator and set it to deliver 0.1 ppm Ozone for at least 30 mins before starting
the calibration, this allow time for the lamp in the ozone generator to stabilise at 0.1 ppm.
Note: The connection to the AQS 1 inlet differs depending on whether you are using the AirCal 1000 or the
Ozone calibrator. See Figure 5-5 and 5-6 to see how the connections differ.
Figure 5-5 When using the Ozone calibrator, make sure the tee fitting is capped
Figure 5-6 When using the AirCal 1000 make sure the tee is un-capped
The NO2 a value is set at the Aeroqual factory and adjusts for the relative response of the O x module
towards ozone calibration gas.
The a value should be checked and adjusted as part of routine field calibration, see Figure 3b.
Note that only the a value is set for NO2, the offset should be left at 0.000 and the gain should be left
at 1.000.
Below is a brief summary of the calibration steps for the NO 2/O3 system. These steps must be carried
out in the order described here. A detailed step by step process is described later in this document.
Calibration steps
Note:
The Ox module gain adjustment is made during the NO2 calibration gas delivery
The NO2 a value adjustment is made during the Ozone gas delivery
The NO2 gain value is NOT adjusted as part of the calibration process
1. Zero air is delivered to the instrument from the AirCal 1000 (or a suitable gas dilution calibrator).
2. The zero offset on the Ox and O3 module is adjusted so that Ox and O3 read zero.
(Confirm NO2 reads zero but do not adjust calibration settings on the NO2 channel).
3. Deliver 0.1 ppm NO2 gas from the calibrator.
4. Adjust Ox gain so the Ox reads 0.1 ppm.
The NO2 module should be reading 0.1 ppm during the 0.1 ppm Ox span calibration. The NO 2 may
read slightly less than 0.1, this is normal, do not adjust the NO 2 gain.
5. Deliver zero air for 10 minutes to allow the sensors to return to baseline.
6. Then deliver 0.1 ppm Ozone.
7. Adjust O3 gain so that the O3 module reads 0.1. The Ox module will be showing a response
towards Ozone during the O3 calibration, this is expected.
If the NO2 shows a response during the Ozone calibration then a new a value must be calculated
according to equation 3 and applied as shown in Figure 5-4.
Set the a value in the Manual Entry dialogue in the Service and Calibration App (Figure 5-4). Check
the NO2 reads zero under these conditions.
7. Deliver Zero air for 10 mins to purge the gas lines before returning to sampling ambient air.
Record details of gas cylinders in Table 3 from the Calibration form in Appendix 5.
Record any changes to the local surroundings which may have an impact on the measurements in
Table 4 from the Calibration form in Appendix 5.
Record details of instrument pre checks in Table 5 from the Calibration form in Appendix 5.
5.6 Record the previously measured flow rate (you should this value recorded in the instrument
journal.
Note: Try to minimise the time the door is open, to keep the sensor readings stable.
5.8 Measure and record the inlet flow rate again after changing the filter.
• If the inlet flow has changed compared to step 5.6,(but less than ± 20 %) since last time it was
measured then check the gas connections for obvious leaks, then use the flow adjustment
valve to correct the flow.
• If the inlet flow has changed (more than ± 20 %) since last time it was measured (step 5.6)
then go to Step 6.
It is important to have a high quality flow meter to perform these flow checks. See Figure 5-2.
6.1. Check all of the individual gas connections for any obvious leaks.
• If a leak is found then correct it, go back to Step 5.9.
6.3 Measure the individual flow rates, then add up the flows to equal total module flow. Record
the sum of the flows in the last column.
• For each module, if the flow rate measured now equals the previously measured flow rate
(± 20 % LPM) then pass for that module, otherwise fail for that module.
• If no specific module is causing a problem and all module flow rates are low then go to 6.4
otherwise go to 6.6.
6.4 Adjust the flow rate using the bypass valve, circle yes if valve adjusted
6.5 Re-measure and record module flow rates for each module.
• If the flow rate measured now equals the previously measured flow rate then pass. Go to
zero calibration.
For those modules which do not have the correct flow rate and cannot be corrected with the
bypass valve then perform a leak check or check for blockages.
• If the problem can be corrected then correct it, otherwise replace the module. Go to zero
calibration.
• If the problem cannot be resolved then remove the module and cap off the gas
connections. Record fail in the calibration report. Go to zero calibration.
Note: that the total flow rate will be reduced because of the module which has
been removed.
It is recommended to leave any failed modules running in the instrument throughout the zero and
span calibration process. After the calibration is complete on the remaining modules, then the door to
the instrument can be opened and the failed modules can be removed for inspection.
The order in which modules should be zero calibrated is:
1) Ox, 2) O3, 3) PID NO2 does not have a zero or gain adjustment applied.
Use the AirCal 1000 to deliver zero air, connect the AQS 1 to the AirCal 1000 as shown in Figure 5-6
with tee open.
7.1.1.Record the current offsets in Table 7, column 7.1.1. Start the zero air flowing. Make
sure there is excess flow out from the tee at the instrument inlet as shown in Figure 5-6
with the tee open.
7.2.2 Wait for the module to stabilise. This means the standard deviation is within the
acceptable range listed in Table 7. Column 7.2.3 (c).
If the values have not stabilised then write fail in the last column in Table 7 column
7.2.7 for this module. Move on to next module and go back to Step 7.2.1.
7.2.3 Record the stabilised reading in Table 7, column 7.2.3 from the 10 minute average.
If the values are within the accepted range, listed in Table 7 column 7.2.3 (b) then no
offset adjustment is required, write pass in column 7.2.7. Move on to next module
and go to Step 7.2.1.
7.2.4 If the stabilised value is outside the acceptable range then an offset adjustment is
required. Calculate and record the new offset using Equation 4.
You can use the calibration app “Recommend” button (Figure 5-3), this will perform the
calculation but you must accept this offset by pushing the “Apply” button, this will upload the
new offset.
7.2.5 If the new calculated offset is within the acceptable range listed in Table 7 column
7.2.4 (b) then upload the new offset.
If the required offset is outside the recommend limits listed in Table 7 do not upload
the offset, write fail in column 7.2.7. Move on to next module and go back to Step
7.2.1.
7.2.6 After 2 or 3 minutes record the readings after the new offset has been uploaded.
Confirm the readings are within acceptable limits.
If yes then pass, if no then fail. Move on to next module and go to Step 7.2.1.
It is recommended to leave any failed modules running in the instrument throughout the zero and
span calibration process. After the calibration is complete on the remaining modules, then the door to
the AQS can be opened and the failed modules can be removed for inspection.
The NO2 calibration, (a adjustment) should be performed at the appropriate step during the Ozone
span gas delivery, Step 8.1.16
The order in which modules should be span calibrated is very important, see section 5.3
8.1.1 Start here for each module. Once all modules have been span calibrated go to Step 9.
Record all initial gains in Table 8 Column 8.1.1
Record the initial NO2 a value in Table 8 Column 8.1.2
8.1.2 Start the NO2 calibration gas flowing at your chosen span point, Aeroqual
recommends 0.1 ppm. Make sure there is excess flow out from the tee at the
instrument inlet as shown in Figure 5-6.
8.1.4 Wait for the Ox module values to stabilise to the NO2 span gas, this means the
readings are within the acceptable noise limits listed in Table 8 Column 8.1.5 (c).
If the values have not stabilised then write fail in the column 8.1.10. Move on to next
module and go to Step 8.1.1.
Note: If the Ox module fails calibration then the NO2 measurement cannot be used.
The O3 module can calibrated even if the Ox module fails calibration.
8.1.5 Record the stabilised Ox values in the calibration form Table 8 Column 8.1.5 from the
10 minute average
If the values are stable and within the accepted range listed in Table 8 Column 8.1.5
(b) then write pass in column 8.1.10. Move on to next module and go to Step 8.1.1.
You can use the calibration app “Recommend” button (Figure 5-3), this will perform the gain
calculation but you must accept this gain by pushing the “Apply” button, this will upload the
new gain.
Note: The calibration app restricts the gain between 0.2 and 5.0. The manual entry screen will
allow the gain to be set outside these limits.
8.1.7 Calculate and record the ratio of the gain change by using Equation 6. Record the
ratio in Table 8, column 8.1.7.
8.1.8 If the calculated gain is within acceptable limits listed in Table 8 Column 8.1.5 (d),
upload the new gain.
If the required gain is outside the recommend limits do not upload the gain, write fail
in the column 8.1.10. Move on to next module and go to Step 8.1.1.
8.1.9 Record the Ox readings after the new Ox gain has been uploaded. Confirm the
readings are within acceptable limits.
If yes then write pass in column 8.1.10, if no then fail. Move on to next module and
go to Step 8.1.1.
8.1.10 Deliver zero air for 10 minutes to purge the gas lines.
Change the gas connection from the AirCal 1000 over to the Ozone calibrator. Make sure to
cap the end of the tee as shown in Figure 5-5.
8.1.11 Deliver ozone from the ozone calibrator at your chosen span point, Aeroqual
recommends 0.1 ppm, make sure the tee is closed using the cap as shown in Figure 5-5.
Note: It takes a while for the ozone delivery to stabilise at the chosen span point, you should
have previously set the output of the ozone generator to your chosen span point and left the
generator running at this span point for at least 30 mins, to allow the output to stabilise.
8.1.13 Wait for the O3 values to stabilise towards O3 calibration gas, as described in Table 8
Column 8.1.5 (c)
If the values have not stabilised then write fail in the last column.
8.1.14 Record the stabilised O3 values in Table 8 Column 8.1.5 from the 10 minute average. If
the values are stable and within the accepted range listed in Table 8 column 8.1.5 (b), then no
gain change is required, then write pass in column 8.1.10.
8.1.15 If the values are stable but outside the acceptable range Calculate and record new O3
gain using Equation 5.
You can use the calibration app “Recommend” button (Figure 5-3), this will perform the gain
calculation but you must accept this gain by pushing the “Apply” button, this will upload the
new gain.
8.1.16 Wait for the Ox values to stabilise towards O3 calibration gas. After the Ox values have
stabilised calculate the a value. The a value is calculated according to this equation:
The a value should be in the range 0.3 to 1.5. If your calculated a value falls outside this
range do not change the a value, contact Aeroqual for technical support:
technical@aeroqual.com.
8.1.17 Upload the a value in the manual entry screen shown in Figure 5-4. Record the new a
value in Table 8 column 8.16b.
8.1.18 After the Ox and O3 gains, and the a value have been applied, record the NO2
readings in Table 8 Column 8.1.9. Do not adjust any calibration settings of the NO2 channel.
8.1.19 Turn off the Ozone gas and return the AQS 1 sampling zero air from the AirCal 1000.
Change the gas connection from the Ozone calibrator over to the AirCal 1000 calibrator.
Make sure to remove the cap from the end of the tee as shown in Figure 5-6.
8.1.19 If your AQS 1 contains a PID gas module, repeat steps 8.1.2 to 8.1.10 for PID.
Follow these post calibration checks and record the results in Table 9 from the Calibration form in
Appendix 5.
9.6 Take AQS 1 out of service mode and record the time in the form.
NOTE: Field calibration is not a substitute for the specified factory service intervals above. Factory
servicing is essential maintenance, and checks the ‘health’ of the internal components.
The K Factor is the applied to the instrument, using the Gain setting for each PM channel:
𝑵𝒆𝒘 𝑮𝒂𝒊𝒏 = 𝑶𝒍𝒅 𝑮𝒂𝒊𝒏 𝑿 𝑲 𝑭𝒂𝒄𝒕𝒐𝒓
You can also perform the adjustment in Microsoft excel or similar.
Field calibration is only as good as the reference data which is used. Using poor quality reference
data for calibration will result in poor quality PM data from your Aeroqual instrument. Aeroqual
recommends that calibration only be carried out using high quality Reference PM Data. If you don’t
have access to a reference instrument, or aren’t confident that you can access a well-run reference
instrument, then you should always rely upon the factory calibration.
Important considerations for field calibration:
• Location of instrument: The Aeroqual instrument and the Reference instrument need to be
located at the same site preferably with their inlets within 10m of each other, and must be
• Instrument set up: Both the Aeroqual instrument and the reference instrument need to be set
up in accordance with the manufacturer’s recommendations, with correctly set flow rates and
fresh filters. If you are using a gravimetric method reference instrument, you must follow
appropriate filter conditioning and weighing methods.
• Duration of colocation: As a general rule, longer is better, as this will give you more data
points and therefore more confidence in your field calibration. It’s important that the colocation
is carried out during the same environmental conditions expected during deployment. 1-2
weeks is preferred and no shorter than 48 hrs.
• Reference instrument selection: Selection of the right reference instrument is critical for a
successful K factor correction. Ideally, reference instruments should be operated in accordance
with local reference method specification. Instrument methods are listed below, in descending
order of preference:
• GRIMM / BAM / TEOM: Are best as they are continuous methods, meaning more data can
be collected in a shorter period of time, giving a more accurate K-Factor. BAM data can be
noisy at hourly averages, so consider using 24hr averages if PM levels are low. This is
Aeroqual’s suggested method, as lots of data can be collected quickly, with no need
for expensive lab work.
• Partisol / Low volume samplers (47mm Filter): Can also be used, however will need to be
run for a longer period of time as they operate on 24hr averages (i.e. 1 data point per
24hrs). Filters must be appropriately prepared and transported. Lab and field blank filters
should also be used. Balance accuracy for pre and post weighing should be better than
±0.0001 µg. High-Volume samplers
• Collect the data from both instruments. Using Microsoft Excel or similar, remove any odd
spikes, error, or automated zero calibration data.
• Select an appropriate averaging period, 24hrs is usually appropriate. Calculate your averages
for both reference and Aeroqual using this formula:
𝑺𝒖𝒎 𝒐𝒇 𝒓𝒆𝒂𝒅𝒊𝒏𝒈𝒔
𝑨𝒗𝒆𝒓𝒂𝒈𝒆 =
𝑵𝒖𝒎𝒃𝒆𝒓 𝒐𝒇 𝒓𝒆𝒂𝒅𝒊𝒏𝒈𝒔
• Using Microsoft excel plot your 2 data sets in an x-y scatter plot. Add a trend line and equation
as below.
• The graph to the left below shows the time series of the two instruments. The graph to the right
below shows the slope calculation. In this example the Dust Sentry is reading higher than the
reference instrument. The slope of the curve Reference vs Dust Monitor is 0.6568. So a gain of
0.6568 needs to be applied to the instrument.
The Aeroqual instrument will need a new gain, calculated using the formula below:
Not all known issues are listed in Table 7-1 especially if not known to Aeroqual when this user guide
was written. The Aeroqual training website provides a comprehensive and up to date list of known
issues and provides solutions as soon as solutions become available.
Online
training.aeroqual.com DM 1.3 Tech training
reference
The Aeroqual technical support team are available to provide assistance when required.
Online
technical@aeroqual.com
reference
When the Aeroqual technical support team are required to help with troubleshooting it is
necessary to see a comprehensive dataset from the AQS 1. This can easily be achieved if the
instrument is online.
If the instrument is not online, use the Manual Sync tool (Section 3.4.1.4) to upload data to
Aeroqual Cloud so the technical support team can provide fast and comprehensive support.
Sensor offline
A sensor offline error is usually caused by either a loose power connection or a loose communications
connection. Check the module connections are secure. If the power connections and the
communications connections are secure then the issue may be inside the module.
Contact the Aeroqual distributor in your local country for technical support.
Communication errors:
1. Failure to see the WIFI network broadcast by the AQS 1
If you cannot see the AQS 1 in your list of WIFI networks it is most likely because the WIFI setting is
set to “Client Mode” and is not broadcasting. See Section 3.2.1.
To resolve this, connect to the AQS 1 through the WIFI LAN and then reset the WIFI mode to “access
point”.
Alternatively the AQS 1 can be reset to factory communication settings by using the USB reset
procedure. See Section 3.3.
If this error occurs contact the Aeroqual distributor in your local country for technical support.
360
483
330 187
More detailed information is in a Technical Note for remote powering the instruments available from
https://www.aeroqual.com/support/technical-documents
Both Mains and Solar power can be used, but these will need to be available and installed
before the instrument arrives When installing power outlets (110 VAC to 230 VAC) for the AQS
1, they should be weatherproof and installed safely. Install 2 power outlets, 1 for the AQS 1 and
one for calibration equipment
Most sites will require a way of securely mounting the AQS 1. Pole Brackets (89 mm / 3 ½ “304
Stainless U-Bolt) are included to mount the instrument on a pole, anchored to the ground. Small
platforms have also been used successfully too. If you have ordered Auxiliary sensors
(Weather, noise etc.) you will need to provide a means of mounting these also.
Mount the instrument to your platform or pole. Assemble the PM inlet and gas inlet, See
Section 2.2.2. Open the door and check that all tubing, power and data connections are
secure.
Check that the Status LEDs on the modules, Embedded PC and Modem (if fitted) are on and
green. Check that the Fan at the rear of the instrument is operating. Check that the compressor
runs.
Put the instrument into Service mode (See Section 4.3.1) and write a comment in the Journal
that you are on site for instrument commissioning.
Check the sampling system for leaks, rectify any leaks (Section 4.4.6 for Particle Monitor and
Particle Profiler, 4.4.10 for Gas)
Check gas module flow rates and adjust to specification if required. Check PM inlet flow rates
and adjust if required.
If you have purchased any auxiliary sensors (Weather, Noise) mount these and run the
cabling through the external glands, and connect to the AUX module. (See Section 2.2.4)
Turn on negative filters to exclude negative values from your dataset.
Write a comment in the Journal that commissioning is complete. Exit Service mode.
Serial number
Customer name
Site name
Engineer name
Ozone generator
Flow meter
Gas Regulator 1
Gas Regulator 2
1 NO2 Air N2
2 Isobutylene Air N2
Pass
Leak check sampling manifold 6.2
Fail
Total
NO2 O3 PID module
flow
7.1.1 7.2.3 7.2.3 (b) 7.2.3(c) 7.2.4 7.2.4 (b) 7.2.5 7.2.6 7.2.7
Max +0.050
O3 0.000 ± 0.005 0.002
Min -0.050
Max +0.050
Ox 0.000 ± 0.005 0.002
Min -0.050
Max +0.05
PID 0.000 ± 0.002 0.002
Min -0.05
Comments:
8.1.1 8.1.2 8.1.3 8.1.5 8.1.5 (b) 8.1.5 (c) 8.1.5 (d) 8.1.6 8.1.6 (b) 8.1.7 8.1.8 8.1.9 8.1.10
Reading after an
NO2 1.000 span ± 5 % 2 % of span NA NA adjustment, (should be
close to 0.000)
max 5.000
O3 N/A span ± 5 % 2 % of span
min 0.2
N/A
max 5.000
Ox N/A span ± 5 % 2 % of span
min 0.2
N/A
max 5.000
PID N/A span ± 5 % 2 % of span
min 0.2
N/A
Comments:
Note:
The Ox module gain adjustment is made during the NO2 calibration gas delivery
The NO2 a value adjustment is made during the Ozone gas delivery
The NO2 gain value is NOT adjusted as part of the calibration process
9 Compliance
1. The Aeroqual AQS 1 Monitor complies with EN 61000-6-1:2001
2. The Aeroqual AQS 1 Monitor complies with EN 61000-6-3:2001
3. The Aeroqual AQS 1 Monitor complies with Part 15 of the FCC Rules.
Operation is subject to the following two conditions: (1) these devices may not cause harmful
interference, and (2) these devices must accept any interference received, including interference that
may cause undesired operation.
NOTE: This equipment has been tested and found to comply with the limits for a Class B digital
device, pursuant to Part 15 of the FCC Rules. These limits are designed to provide reasonable
protection against harmful interference in a residential installation. This equipment generates, uses
and can radiate radio frequency energy and, if not installed and used in accordance with the
instructions, may cause harmful interference to radio communications. However, there is no
guarantee that interference will not occur in a particular installation. If this equipment does cause
harmful interference to radio or television reception, which can be determined by turning the
equipment off and on, the user is encouraged to try to correct the interference by one or more of the
following measures:
10 Warranty
The AQS 1 warranty policy can be found on the main company website:
www.aeroqual.com/support/warranty