User's Manual: Engine Ignition Analyzer
User's Manual: Engine Ignition Analyzer
User's Manual: Engine Ignition Analyzer
Page 2
3. GENERAL DESCRIPTION
3.1 Instrument description
LCD display
Protective hoster
MODE button
20
10
Measurement
display area 0
-10
-4 -2 0 2 ms
• When the instrument is OFF, to turn it ON press and hold the ‘POWER/CLEAR’
button until the unit turns on (in approximately 1 second).
• When the instrument is ON, to turn it OFF press and hold the “POWER/CLEAR’
button until the display turns OFF (in approximately 3 seconds).
• When the instrument is ON, press the ‘POWER/CLEAR’ button to clear all
measurement data, and start a new measurement. This operation can be also
be performed to re-scale the measurements and optimize viewing in the display.
• The ‘Auto Power Off’ feature will automatically turn the instrument off after 3
minutes of no button being pressed or 15 seconds after the last spark signal was
detected. Turning the unit off manually when not in use will prolong battery life.
01607 RPM
4 4 Stroke
2 2 Stroke
Waste Spark
0
-5
-4 -2 0 2 ms
Page 4
4.3 ‘MODE’ button and menu
The ‘MODE’ button allows the selection of the type of measurement to be
shown in the display. Upon pressing this button a pull down menu will open with
the current setting highlighted, To change the setting press the ‘MODE’ button
repeatedly until the desired setting is highlighted, then wait until the pull down
menu closes. The new current setting will be displayed as the mode icon at the
top of the screen.
01607 RPM
kv
Burn Time
10
Dwell
5
Tachometer
0
Voltage
-5
-4 -2 0 2 ms
The ’VIEW’ button is used to select how the chosen measurement will be
shown in the display. Upon pressing this button a pull down menu will open
with the current setting highlighted. To change the setting press the ‘VIEW’
button repeatedly until the desired setting is highlighted, then wait until
the pull down menu closes. The new setting will be displayed as the display
icon at the top of the screen.
Page 5
01607 RPM
kv
Chart
20
Comparison
10
Digital
0
Waveform
-10
-4 -2 0 2 ms
Comparison view.
Full Empty
NOTE: The instrument will turn automatically off if the battery voltage falls below the
absolute minimum.
Page 6
5. MEASURING MODES
There are four different ignition system parameters that can be selected for measurement,
depending on the engine’s ignition system:
• Engine RPM.
• Spark burn (firing) time.
• Spark plug voltage.
• Dwell angles (for ignition systems equipped with high voltage spark plug wires).
• Primary ignition coil current ramp time (for coil on plug and coil near plug ignition systems).
5.4 Dwell angle (for ignition systems using spark plug wires only)
Dwell angle is a measurement of the angle of rotation of the crankshaft, between the moment
in which the primary of the ignition coil is energized (e.g. point contacts closed) and the
spark is generated (e.g. point contacts opened). Its main use is in determining if there is
enough angle (or time) to energize the primary of the ignition, and not so much as to cause
the ignition coil to overheat and fail. Dwell angles are specified for a particular engine and
RPM, and usually adjusted in the mechanical distributor/point assembly.
5.5 Current ramp time (for coil on plug and coil near plug only)
Ramp time is measured from when the primary coil current first begins to increase to
when it stops increasing. The current may stop increasing because the coil has reached
its saturation point, or it is interrupted to generate a spark.
Ramp time measurements are useful for diagnosing problems in the primary circuit of
the ignition module. Abnormally long ramp times may indicate excessive primary circuit
resistance, low input voltage, that the wrong module is installed, etc. Shorter than normal
ramp times may indicate a shorted primary coil, that the wrong module is installed, etc.
Page 7
6. DISPLAY VIEWS
There are four views that can be selected to display measurements:
• Chart
• Comparison
• Digital
• Waveform
01607 RPM
RPM
0
S
• ‘MODE’ button:
Pressing the ‘MODE’ button once will open the mode menu, and highlight the measurement
mode currently selected; pressing the ‘MODE’ button again while the menu is open will
select the next available measurement mode. All four measurements (spark burn time,
RPM, spark plug peak voltage, and dwell angle or ramp time) are simultaneously tracked
and stored, and are available for viewing at any time (without delay).
• ‘HOLD’ button:
Pressing this button will pause the measurement, and hold the display in its the current
condition. Pressing the ‘HOLD’ button again will resume normal measurement.
15 Maximum
10 Average
5 Minimum
01607 RPM
8
6 10 7.14MAX
4 12 5.20MIN
6.71
PEAK
2
0 kV
Fig. 8 - Digital mode display
Page 9
• ‘POWER/CLEAR’ button:
Pressing the ‘POWER/CLEAR’ button clears all digital readout values, and re-scales
the gauge if necessary.
• ‘MODE’ button:
Pressing the ‘MODE’ button once will open the mode menu, and highlight the measurement
mode currently selected; pressing the ‘MODE’ button again while the menu is open will
select the next available measurement mode. All four measurements (spark burn time,
RPM, spark plug peak voltage and dwell angle and ramp time) are simultaneously tracked
and stored, and are available for viewing at any time (without delay).
• ‘HOLD’ button:
Pressing this button will pause measurement, and hold the display in the current
condition. Pressing the ‘HOLD’ button again will resume normal measurements.
6.4 Waveform view
This mode displays the signal amplitude over time (like an oscilloscope). The scale and
units of the measurement are located on the vertical axis, at the far left of the display. The
time scale and units are shown below the horizontal axis, at the bottom of the screen.
01607 RPM
kv
20
10
-10
-4 -2 0 2 ms
Flexible probe
Extension cable
1- Connect the extension cable to the instrument using their BNC connectors.
2- Insert and firmly press the appropriate COP or clip-on SPW sensor into the barrel
connector of the extension cable.
7.3 Measuring coil on plug and coil near plug ignition systems
1- Ensure the instrument is turned off.
2- Insert the COP sensor barrel connector plug into the flexible probe receptacle, removing
the SPW sensor if necessary.
Page 11
3- Turn the instrument on and select the appropriate cycle for the engine under
measurement.
4- Select the desired mode and view.
5- Place the COP sensor on the top and center of the ignition module (as shown in the
Fig. 12), and observe whether the instrument is able to detect and display the desired
measurement.
6- If the instrument is unable to detect the ignition system signal, it may be necessary to
reposition the sensor so that a consistent signal is detected and displayed.
NOTES
• The position of the sensor and the particular design of the ignition module may
affect the signal’s shape shown when the waveform display is selected. In order to
compare waveforms between several cylinders’ ignition modules, the placement of the
sensor relative to the ignition module should be kept as consistent as possible for all the
measurements.
• In some engines, the ignition modules are located very close together, and in rare
occasions this could cause the instrument’s sensor to detect signals from nearby modules,
leading to inconsistent measurements. This can usually be solved by changing the
placement or orientation of the sensor on the ignition module.
7.4 Measuring ignition systems with (high voltage) spark plug wires
1- Ensure the instrument is turned off.
2- Insert the SPW or Clip-on SPW sensor barrel connector plug into the flexible probe or
extension cable receptacle.
3- Turn the instrument on, and select the appropriate cycle for the engine under
measurement.
4- Select the desired mode and view.
5- If using the SPW sensor, place the spark plug wire into the slot of the sensor, and as far
as possible from other spark plug wires trying to maintain a right angle (90°) between
the sensor and the wire.
7- If the instrument is unable to detect the ignition system signal, it may be necessary to
reposition the sensor so that a consistent signal is detected and displayed.
NOTES
• The absolute spark plug voltage reading will depend on the position of the sensor, the
particular characteristics of the wire, etc.. Therefore the placement of the sensor relative
to the wire should be keep as consistent as possible for all the measurements in order
to compare voltage readings between several cylinders’ spark plug wires.
• Spark burn time, dwell angle and RPM measurements are not sensitive to the exact
position of the sensor relative to the spark plug wire, but the sensor must be positioned
so the instrument is capable of detecting the signal.
• When several spark plug wires are routed or bunched close together, the
capacitive sensor may receive signals from two or more wires simultaneously, which
could cause erroneous measurements. In these situations it is necessary to separate
the wire under measurement from the others, in order to reduce interference and obtain
an accurate measurement.
8. RECHARGING THE INSTRUMENT
1- If still operating, turn the instrument power off.
2- Locate and lift the rubber cap (attached to the protective holster and located besides
the BNC connector) to reveal the micro USB receptacle.
Page 13
3- Insert the USB charger cable’s micro USB plug into the receptacle of the instrument
matching the orientation of the connectors.
4- Insert the USB type A connector of the cable into the USB port of the supplied USB
power charger adapter, and plug the power adapter into an AC power outlet.
5- Upon start of charging the battery, the display will show the charging in progress icon
and the screen will turn off after a few seconds.
6- If the instrument is turned on (by pressing the ‘POWER/CLEAR’ button) at any time
while connected to the power adapter, the instrument will display the charging status .
7- The USB type A connector can be also plugged into a USB port of a personal computer,
powered USB hub, or any other USB compliant power source.
5- The recharge time will depend on the state of charge of the battery, and it may take up
to 6 hours to fully recharge a depleted battery.
NOTES
• If the instrument has not been used for a long time, or the state of charge of the
battery has fallen to a critically low level, a period of pre-conditioning of the battery is
automatically added to the normal charging cycle, which increases the charging time
in order to restore the battery capacity to its maximum, No user intervention is needed,
and this process is automatically carried out by the instrument.
• If the battery is completely depleted, pressing the power on button will not turn the
instrument on, and it must be recharged before it can be used again.
9. MAINTENANCE
Keep the instrument in its carrying case when not in use and do not subject it to dampness,
severe heat or cold. Do not use the instrument in the rain; if it should accidentally get wet,
dry it off with a clean paper towel before storing it away.
Protect the unit from contact with any solvents. Never clean with a solvent or petroleum
based medium such as gasoline, as these chemicals may attack the plastic parts and
cause permanent damage. Never use an abrasive cleaner. Cleaning should be limited to
wiping with a clean damp paper towel and a small amount of soap if required. Dry the unit
thoroughly after any cleaning.
The unit is a sealed instrument, and contains no user serviceable parts. Opening this
instrument will void the warranty.
10. WARRANTY
This instrument carries a one (1) year warranty (from the date of purchase by the original
owner) against defects of material or workmanship. For details see our Standard Warranty
Information on our web page at www.gtc.ca, or request a printed copy.
Page 14
Part # GTC505MN1707EN
To visit our website, scan the QR code below with your smart phone.
www.gtc.ca
“SmarTach”, “V Ready”, “GTC505” and “GTC” are trademarks of General Technologies Corp.